Instruction/ maintenance manual of the product FCLTX-100 Omega
Go to page of 6
PRODUCT INSTRUCTION SHEET Percent of free chlorine in protonated form (HClO) 0 10 20 30 40 50 60 70 80 90 100 5 5.5 6 6.5 7 7.5 8 8.5 9 9.5 10 pH percent SECTION 1.
SECTION 3.0 SENSOR PREP ARA TION FIG. 2 3.0 FREE CHL ORINE SENSOR ASSEMBL Y . The Free Chlorine Sensor is shipped with the membrane cap pre-installed and cov ered with a cap with water inside to keep the membrane wet. Make sure to keep sensor cap, anode and cathode inside the sensor body , away oily or gr easy materials.
Po wer Supply PL C, DVM,... sensor red wire sensor black wire FIG. 3 PRODUCT INSTRUCTION SHEET 6.0 ELECTRICAL INST ALLA TION. The sensor produces an ap- proximate output of 4 mA in air and 20mA at the t op range of free chlorine output (0-2ppm, 0-5ppm, 0-10ppm).
ZERO REA D ClO 2 C L O 2 to Increase output to Decrease output SECTION 8.0 CALIBRA TION FIG. 6 SECTION 9.0 SENSOR ST ORAGE IMPORT ANT NOTE: SENSORS ARE SUPPLIED F ACT ORY CALIBRA TED WITH A 4-20mA SIGNAL OUTPUT CORRESPONDING TO THEIR SPECIFIC RANGE (0-2, 0-5 OR 0-10ppm).
PRODUCT INSTRUCTION SHEET 10.1 GENERAL MAINTENANCE. Ser vice the sensor regularly (weekly is suggested) 10.2 MEMBRANE CAP REPLACEMENT . If membrane replace- ment is required , a new cap with preinstalled membrane must be used. Order FCLA-2016 replacement membrane cap and FCLA-2015 refill solution.
PRODUCT INSTRUCTION SHEET Operating temperature range: 0-45 degC Maximum operating pressure: 1 bar/14.7 psi/1atm Flow ra te minimum: .20 gpm Flow ra te maximum: 0.
An important point after buying a device Omega FCLTX-100 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Omega FCLTX-100 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Omega FCLTX-100 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Omega FCLTX-100 you will learn all the available features of the product, as well as information on its operation. The information that you get Omega FCLTX-100 will certainly help you make a decision on the purchase.
If you already are a holder of Omega FCLTX-100, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Omega FCLTX-100.
However, one of the most important roles played by the user manual is to help in solving problems with Omega FCLTX-100. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Omega FCLTX-100 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center