CL23A, 24, 25, 26, 27 Digital Calibrator/Thermometer

Page 2 of the Omega 24 user manual — For immediate technical or application assistance:
Page 2 · click to enlarge
Page 2

omega.com

TM

OMEGA

OMEGAnet SM On-Line Service Internet e-mail http://www.omega.com [email protected]

Servicing North America: USA: One Omega Drive, Box 4047 ISO 9001 Certified Stamford, CT 06907-0047 Tel: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com Canada: 976 Bergar Laval (Quebec) H7L 5A1 Tel: (514) 856-6928 FAX: (514) 856-6886 e-mail: canada@omega.com

For immediate technical or application assistance:

USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGASM Customer Service: 1-800-622-2378 / 1-800-622-BESTSM Engineering Service: 1-800-872-9436 / 1-800-USA-WHENSM TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA Mexico and Latin America: Tel: (95) 800-TC-OMEGASM FAX: (95) 203-359-7807 En Espan˜ol: (203) 359-1660 ext: 2203 e-mail: espanol@omega.com

Servicing Europe: Benelux: Postbus 8034, 1180 LA Amstelveen, The Netherlands Tel: (31) 20 6418405 FAX: (31) 20 6434643 Toll Free in Benelux: 06 0993344 e-mail: nl@omega.com Czech Republic: Ostravska 767, 733 01 Karvina Tel: 42 (69) 6311899 FAX: 42 (69) 6311114 e-mail: czech@omega.com France: 9, rue Denis Papin, 78190 Trappes Tel: (33) 130-621-400 FAX: (33) 130-699-120 Toll Free in France: 0800-4-06342 e-mail: france@omega.com Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany Tel: 49 (07056) 3017 FAX: 49 (07056) 8540 Toll Free in Germany: 0130 11 21 66 e-mail: germany@omega.com United Kingdom: 25 Swannington Road, P.O. Box 7, Omega Drive, ISO 9002 Certified Broughton Astley, Leicestershire, Irlam, Manchester, LE9 6TU, England M44 5EX, England Tel: 44 (1455) 285520 Tel: 44 (161) 777-6611 FAX: 44 (1455) 283912 FAX: 44 (161) 777-6622 Toll Free in England: 0800-488-488

e-mailuk@ omega.com

It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.

The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, patient connected applications.

®

Page 3 of the Omega 24 user manual — omega.com
Page 3 · click to enlarge
Page 3
Page 3 of the Omega 24 user manual — omega.com
Page 4 of the Omega 24 user manual — COMMON SPECIFICATIONS
Page 4 · click to enlarge
Page 4

GENERAL INFORMATION This manual provides operating instructions and maintenance information for five temperature instruments. These instruments are high performance cali- brator-thermometers capable of simulating and measuring a wide-variety of sensors. In addition, features such as data storage and ramping further enhance their versatility.

It is recommended that you read this manual thoroughly, especially the sec- tions on safety, prior to operating these instruments.

COMMON SPECIFICATIONS

THERMOMETER INPUTS: THERMOCOUPLE; Miniature TC jack. RTD, OHMS, THERMISTOR; T-series instrumentation connector.

CALIBRATOR OUTPUTS: THERMOCOUPLE; Miniature TC jack. RTD, OHMS, THERMISTOR; T-series instrumentation connector.

THERMOCOUPLE TYPES: K, J, T, E, N, B, R, S, G, C, D.

RTD TYPES1 100 Ω platinum ( ∝ = 0.00385)

1000Ωplatinum (∝= 0.00385)

THERMISTOR TYPE: 2252Ω(YSI Series 400).

RANGES: RTK; –200°C to 1372°C, –328°F to 2502°F RTJ; –210°C to 760°C, –346°F to 1400°F RTT; –200°C to 400°C, –328°F to 752°F RTE; –230°C to 1000°C, –382°F to 1832°F RTN; –200°C to 1300°C, –328°F to 2372°F RTB; 500°C to 1820°C, 932°F to 3308°F RTR; 0°C to 1768°C, 32°F to 3214°F RTS; 0°C to 1768°C, 32°F to 3214°F RTG; 300°C to 2316°C, 572°F to 4201°F RTC; 0°C to 2316°C, 32°F to 4201°F RTD; 0°C to 2316°C, 32°F to 4201°F RTD; –200°C to 850°C, –328°F to 1562°F THERMISTOR; –40°C to 150°C, –40°F to 302°F

OHMS; 0-999.99Ω OHMS; 0-9,999.9Ω OHMS; 0-99,999Ω

ACCURACY (18°C to 28°C ambient, 2 years)*:

K, J, T, E, N; ± 0.5°F (rdg ≥–50°F) K, J, T, E, N; ±1.0°F (rdg < –50°F) B, R, S, G, C, D; ±1.7°F

Page 5 of the Omega 24 user manual — CALIBRATOR-MODE EXCITATION-CURRENTS:
Page 5 · click to enlarge
Page 5

RTD; ±0.2°F (rdg ≥– 50°F) RTD; ±0.5°F (rdg < – 50°F) RTD; ±0.2°F ±0.04% rdg (meter mode) THERMISTOR; ±0.5F

OHMS; ±0.02% range

* Exclusive of sensor errors, and lead resistance induced errors.

RESOLUTIONTEMPERATURE; 0.1°/1° F/C RESOLUTION: OHMS; 0.01 Ω /0.1 Ω /1 Ω

REPEATABILITY (1 week at constant ambient temperature):

K, J, T, E, N; ±0.2°F typ. B, R, S, G, C, D; ±1.0°F typ. RTD, THERMISTOR; ±0.1°F typ.

OHMS; ±0.01% rng typ.

TEMPERATURE COEFFICIENT:

From 18°C to 28°C, included in accuracy specifications. From 0°C to 18°C, and 28°C to 50°C:

K, J, T, E, N; ±(0.002% rdg + 0.03°C)/°C B, R, S, G, C, D; ±(0.002% rdg + 0.1°C)/°C RTD, THERMISTOR, OHMS; ±(0.002% rdg + 0.004% rng)/°C

METER-MODE INPUT-CURRENT (THERMOCOUPLES): 50nA typ.

METER-MODE READING-RATE: 1 reading/second typ.

METER-MODE EXCITATION-CURRENTS:

100ΩRTD;

CALIBRATOR MODE OUTPUT LOAD (THERMOCOUPLES): 500Ωmin.

CALIBRATOR-MODE EXCITATION-CURRENTS:

100ΩRTD; 0.5 – 1mA (0.1 – 2mA typ.) 1000ΩRTD; 50 – 100µA (10-200µA typ.) 999.99Ω; Same as 100ΩRTD 9999.9Ω; Same as 1000ΩRTD 99999Ω; 10 – 20µA (1 – 20µA typ.) THERMISTOR; T ≤113°F, 10 – 20µA

T> 113°F, 0.5 – 1mA

MAXIMUM COMMON MODE VOLTAGE: 42V peak to earth.

T> 113°F,

Page 6 of the Omega 24 user manual — TEMPERATURE COEFFICIENT:
Page 6 · click to enlarge
Page 6
KEYPAD12 momentary switches with tactile feedback select;
  • On/Off
  • Change/Enter
  • Clear/–0
  • Sensor Select/1
  • °F/°C/2
  • Resolution/3
  • Operate/4
  • Store/5
  • Recall/6
  • Calibrator/Meter/Up Ramp/7
  • Calibrator Step/Down Ramp/8
  • Calibrator Ramp/Meter Hold/9
DISPLAY5 digit LCD, 0.4" height, and decimal point. Annunciators;
  • Loaded Memory Locations (10 for data storage, 4 for ramp parameters)
  • Low Battery (BAT)
  • Number Change Mode (NUM)
  • Calibrator Mode (CALIB)
  • Meter Mode (METER)
  • Operate Mode (OPR)
  • Store Mode (STO)
  • Recall Mode (RCL)
  • Meter Hold (HOLD)
  • Scale (°F/°C)
  • Sensor Type
POWER OFF CONFIGURATION RETENTIONInstrument retains last selected;
  • Sensor Type
  • °F/°C Scale
  • 0.1°/1° Resolution
  • Calibrator/Meter Mode
  • Memory Contents (data storage and ramp parameters).
DIAGNOSTICSDisplay codes indicate following conditions;
  • 'BAT': Low Battery
  • 'OPEN': Open Input Thermocouple, RTD or Thermistor.
  • 'E-1' (momentary): Invalid Keypad Entry
  • 'E-2' (momentary): Ramp Function Not Installed
  • 'HI': Temperature Reading or Keypad Entry Exceeds Rating
  • 'LO': Temperature Reading or Keypad Entry Too Low For Accuracy
  • LCD Test: During power-up, all segments and annunciators enabled momentarily

ENVIRONMENTAL LIMITS FOR OPERATING: 0°C to 50°C, less than 90% relative humidity (R.H.) up to 35°C; reduce R.H. limit by 3%/°C from 35°C to 50°C.

ENVIRONMENTAL LIMITS FOR STORAGE: –35°C to 60°C, less than 95% relative humidity (R.H.) up to 35°C; reduce R.H. limit by 3%/°C from 35°C to 60°C.

POWER: 9 volt alkaline battery (NEDA 1604A).

Page 7 of the Omega 24 user manual — UNIQUE SPECIFICATIONS
Page 7 · click to enlarge
Page 7

BATTERY LIFE, CONTINUOUS: 10 hrs. typical, alkaline; 3 hrs typical, Ni-Cd (rechargeable).

BATTERY INDICATOR: Display indicates BAT when less than 10% of life remains.

DIMENSIONS, WEIGHT: 7.0" × 2.9 × 1.1". Net weight 12 oz.

UNIQUE SPECIFICATIONS

  • MODEL CL23A/CL23EC: SENSOR SELECTION: K, J, T/K, J, T, L Thermocouples

ACCURACY: (18°C to 28°C ambient, 2 years): ±0.5°F (rdg ≥–50°F).

±0.04% rdg (rdg > 1250°F) ±1.0°F (rdg < –50°F)

TEMPERATURE COEFFICIENT: 18°C to 28°C; included in accuracy specifications. From 0°C to 18°C, and 28°C to 50°C; less than ±(0.006% rdg +0.03°C)/°C.

RAMP FUNCTION: Not Included.

BATTERY LIFE, CONTINUOUS: 16 hrs typical, alkaline; 4 hrs typical, Ni-Cd (rechargeable).

  • MODEL CL24: SENSOR SELECTION: K, J, T, E Thermocouples

ACCURACY: (18°C to 28°C ambient, 2 years): ±0.5°F (rdg ≥–50°F).

±0.04% rdg (rdg > 1250°F) ±1.0°F (rdg < –50°F)

TEMPERATURE COEFFICIENT: 18°C to 28°C; included in accuracy specifications. From 0°C to 18°C, and 28°C to 50°C; less than ±(0.006% rdg +0.03°C)/°C.

RAMP FUNCTION: Not Included.

BATTERY LIFE, CONTINUOUS: 16 hrs typical, alkaline; 4 hrs typical, Ni-Cd (rechargeable).

  • MODEL CL25/CL25EC: SENSOR SELECTION: K, J, T, E, N, B, R, S, G, C, D/K, J, T, E, N, B, R, S, L, C, D Thermocouples

RAMP FUNCTION: Included.

BATTERY LIFE, CONTINUOUS16 hrs typical, alkaline; 4 hrs typical, Ni-Cd (rechargeable)
  • MODEL CL26: SENSOR SELECTION: K, J, T, E Thermocouples

100ΩPlatinum RTD 2252ΩThermistor 1,000Ω/100,000ΩResistance

Page 8 of the Omega 24 user manual — UNIQUE SPECIFICATIONS
Page 8 · click to enlarge
Page 8

ACCURACY: (18°C to 28°C ambient, 2 years): ±0.5 °F (rdg ≥–50°F).

±0.04% rdg (rdg > 1250°F) ±1.0°F (rdg < –50°F) OHMS; ±0.04% range

TEMPERATURE COEFFICIENT: 18°C to 28°C; included in accuracy specifications. From 0°C to 18°C, and 28°C to 50°C.

K, J, T, E; ±(0.006% rdg + 0.03°C)/°C RTD, THERMISTOR, OHMS; ±(0.006% rdg +0.012% rng)/°C

RAMP FUNCTION: Not Included.

  • MODEL CL27/CL27EC : SENSOR SELECTION: K, J, T, E, N, B, R, S, G, C, D/K, J, T, E, N, B, R, S, L, C,

D Thermocouples 100Ω/1000ΩPlatinum RTD 1,000Ω/10,000ΩResistance

RAMP FUNCTION: Included.

  • FEATURES
  • Combination calibrator, thermometer and datalogger.
  • Function selection and numeric data entered via 12 keyswitch color coded keypad.
  • Keyswitches have tactile feedback.
  • 5 digit LCD includes annunciators for operating modes.
  • Up to 11 thermocouple types, 2 RTD types, and thermistor.
  • 1° or 0.1° resolution over temperature range of each sensor.
  • °F and °C scales.
  • Reading hold mode.
  • 10 memory locations for saving meter readings and/or calibrator outputs.
  • Step mode for quick sequential recall of calibrator outputs from memory.
  • Manual/automatic ramping for testing controller setpoints.
  • Conforms to ITS-90 thermocouple tables.
  • Dust proof, splash proof, drop proof construction.
  • Built-in tilt stand/hanger for bench use or hands free field measurements.
  • User friendly programming.
  • Retains data and programming, even when turned off.
  • Performs diagnostic tests and indicates fault conditions.
  • Low battery and open sensor indications.

MANUAL ADDENDA Improvements or changes to this manual will be explained on an addendum included with the instrument. All change information should be incorporated immediately into the appropriate places in the manual.

Page 9 of the Omega 24 user manual — WARNING
Page 9 · click to enlarge
Page 9

UNPACKING YOUR UNIT Remove the Packing List and verify that you have received all equipment. If you have any questions about the shipment, please call the OMEGA Customer Service Department at 1-800-622-2378 or (203) 359-1660.

When you receive the shipment, inspect the container and equipment for any signs of damage. Note any evidence of rough handling in transit. Immediately report any damage to the shipping agent.

Make sure that you have the following items in the packing box.

  • 1 CL20 Series Calibrator and Holster.
  • 1 Operator’s Manual
  • 1 NIST Certificate
  • 1 Accessories Kit

SAFETY INFORMATION

SAFETY SYMBOLS AND TERMS

The symbol on the instrument denotes that the user should refer to the

L !

operating instructions.

The WARNING used in this manual explains dangers that could result in

personal injury or death.

The CAUTION used in this manual explains hazards that could damage

the instrument.

SAFETY PRECAUTIONS

WARNING

These instruments are intended for use by qualified person- nel trained in the safe operation of electronic testing equip- ment. Read the instruction manual thoroughly before using, to become familiar with the instrument's operations and capabilities.

Page 10 of the Omega 24 user manual — CAUTION
Page 10 · click to enlarge
Page 10

WARNING

Do not touch a temperature probe sheath when measuring excessively high or low temperatures, or toxic substances.

WARNING

Do not attempt to measure temperatures beyond the range of the probe being used. Probe damage or personal injury could result from exceeding a probe's maximum temperature rating.

WARNING

The American National Safety Institute (ANSI) states that a shock hazard exists when probes or sensors are exposed to voltage levels greater than 42VDC or 42V peak AC. Do not use this instrument where voltages at the measurement sur- face exceed these levels.

WARNING

Do not substitute a metal part for the nylon screw in the rear case. Doing so will degrade electrical isolation of the case.

WARNING

The battery is accessible through a cover on the back of the instrument. To avoid electrical shock hazard, disconnect all temperature probes and sensors and turn the unit off before removing the cover.

WARNING

Never use this instrument or any probe or sensor inside a microwave oven.

CAUTION

Avoid making sharp bends in probe or sensor lead wires. Bending lead wires at a sharp angle can damage the wire causing probe failure.

Page 11 of the Omega 24 user manual — WARNING
Page 11 · click to enlarge
Page 11

CAUTION

Keep inputs electrically isolated from outputs. Do not con- nect an external voltage source between the two connectors, and avoid any external electrical pathways between input and output (T2 & T1).

Failure to do so will disturb meter readings and calibrator output levels. Damage to the instrument is also possible.

Where the meter and calibrator connectors are simultane- ously wired into a system lacking the above isolation, it is necessary to use an "ungrounded" (i.e. electrically isolated) probe at the thermometer input.

BATTERY INSTALLATION/REPLACEMENT A 9V alkaline battery is supplied with the instrument but not installed. Read the following installation instructions before attempting to install or remove the battery. Replacement batteries should also be alkaline types.

WARNING

Turn the unit off and disconnect any input/output connec- tions before replacing the battery. Put the cover back into place on the battery compartment before resuming use of the instrument.

1. Remove the cover from the battery compartment by sliding it off in the direction of the arrow located on the battery cover.
Omega 24 — WARNING diagram (page 11)
Fig. 3

9 Figure 1. Battery Installation

Page 12 of the Omega 24 user manual — OPERATION WITH RECHARGEABLE
Page 12 · click to enlarge
Page 12
2. Remove the old battery by lifting the battery extractor loop.
3. Place the new battery in the battery compartment. Be sure to observe proper polarity.
4. Re-install the battery cover before resuming use of the instrument.
  • NOTES:
  • Less than 10% of battery life remains when the BAT annunciator turns on.
  • When the battery is excessively discharged, the display is blanked.
  • If the instrument is going to be stored for a long period of time or in a high temperature environment, remove the battery to prevent leakage damage.
  • After a new battery is installed, allow approximately 5 seconds for display turn-on the first time the unit is turned on. During this period, microcomputer initialization is performed.

MEMORY BACKUP During battery replacement, the contents of user-programmed memory (data, operating modes, etc.) can be saved. Prior to removing the old battery, turn off the instrument, and connect an HH22-AC battery charger. Then exchange batteries, and disconnect the battery charger. Do not leave the bat- tery charger connected to instruments with non-rechargeable batteries.

OPERATION WITH RECHARGEABLE

BATTERY

HH22-AC provides a 9-volt Ni-Cd battery and recharger suitable for use with the unit. This battery provides 3-4 hours of continuous operation. This dura- tion can be extended to several days by operating simultaneously from both the battery and recharger.

Turn off the instrument to recharge the battery. Allow 7 hours to fully recharge the battery. Then unplug the recharger to avoid overcharging.

Page 13 of the Omega 24 user manual — 2. INITIAL TURN ON
Page 13 · click to enlarge
Page 13

OPERATING INSTRUCTIONS

1. DIAGNOSTICS AND ERROR MESSAGES

Unit self-diagnostics provide fault condition readouts which are described below:

PROBLEM: LIKELY CAUSE:

Blank display, unit does not (1) Improper battery installation. power-up. (1) Check battery polarity. (2) Dead battery.

BAT Annunciator. (1) Low battery voltage, install a new

battery. If problem persists, con- sult factory.

Display reads OPEN. (1) No thermocouple, RTD or thermis-

tor is plugged into the meter input.

Display reads E1 momentarily. (1) This indicates that an invalid key-

pad entry has been made. Review keystroke sequence, or consult manual for input instructions.

Display reads E2 momentarily. (1) Attempt made to activate non-

installed ramp function. NOTE: RAMP Function enabled on Models CL25 and CL27 only.

Display reads HI. (1) Meter-mode input-temperature exceeds rating of selected sensor. (2) Calibrator-mode keypad-entry exceeds rating of sensor type.

Display reads LO. (1) Meter-mode input-temperature too

low for accurate measurement. (2) Calibrator-mode keypad-entry too

low for accurate simulation.

2. INITIAL TURN ON

Turn on the instrument by depressing the ON-OFF keyswitch once (Figure 2). At power-up, the unit first performs a display test. All segments and annunciators of the liquid crystal (LCD) are momentarily turned on for visual confirmation by the user (Figure 3).

factory.

Page 14 of the Omega 24 user manual — OPERATING INSTRUCTIONS
Page 14 · click to enlarge
Page 14
Omega 24 — OPERATING INSTRUCTIONS diagram (page 14)
Fig. 4

Figure 2. Keypad

Omega 24 — OPERATING INSTRUCTIONS diagram (page 14)
Fig. 5

Figure 3. Display Test

If the unit is turning on for the first time after a new battery is installed, it will automatically default to the METER mode (ie. temperature measurement operation). It also defaults to K-type thermocouples, and °F readings with 0.1° resolution. The corresponding LCD annunciators are enabled:

Page 15 of the Omega 24 user manual — OPERATING INSTRUCTIONS
Page 15 · click to enlarge
Page 15

METER: (unit functioning as a thermometer) K: (K-type thermocouple) °F: (Fahrenheit readings) OPR: (Instrument is operational)

If no input thermocouple is connected to the unit at this time, the numeric segments will indicate "OPEN".

Plugging a thermocouple into the meter jack (connector T2, on the top right side) will give actual thermocouple temperature readings.

If the instrument is to be used as a thermometer or ohmmeter, refer to the Meter Mode Operation section for detailed instructions. For use as a tem- perature-calibrator or ohms simulator, refer to Calibrator Mode Operation.

3. METER MODE OPERATION (Thermocouples)

Set-Up:

1. Turn on instrument by depressing the ON/OFF key.
2. Check that the BAT annunciator turns off following the power-up LCD test. If it does not turn off, less than 10% of battery remains. Refer to Battery Installation/Replacement instructions.
3. If the meter is already in the METER mode, the METER annunciator will be on. Otherwise the calibrator mode is enabled and the CALIB annunciator is on. Depress the METER key to return to meter mode.
4. Connect an input thermocouple (TC) to connector T2. Select the appropriate TC Type with the select-key. A display annunciator will indicate the selected TC-type.

NOTE: Space and legibility constraints on the display preclude the use of dedicated annunciators for the less common tungsten- rhenium thermocouple types (G, C & D). Instead, these types are annunciated momentarily on the alpha-numeric display when first selected. Both T and R annunciators remain on to indicate a Tungsten-Rhenium selection. To recall the actual G, C, D selec- tion, turn the unit off, then on again for another momentary read- out.

5. Select temperature scale with °F/°C key. The corresponding annunciator is enabled.
Page 16 of the Omega 24 user manual — OPERATING INSTRUCTIONS
Page 16 · click to enlarge
Page 16
6. Select display resolution with 0.1°/1° key.

Hold: Meter readings can be put on hold at any time by depressing the HOLD key. At this time, the HOLD annunciator turns on, the reading is frozen, and all keys (except ON/OFF and HOLD) are locked out. Depress the HOLD key again to return to real-time readings, and re-enable the keypad.

4. METER MODE OPERATION (RTD, Thermistor & Ohms)

Set-Up:

1. Connect a sensor or unknown resistance to the instrument as shown in Figures 4, 5, or 6.

The 2-wire configuration is simplest, but includes lead-wire resistance in the measurement. 2-wire measurements are generally limited to high resistance sensors (e.g. 1000-ohm RTDs, 2252-ohm thermistors).

Omega 24 — OPERATING INSTRUCTIONS diagram (page 16)
Fig. 6

Figure 4. 2-Wire Resistance Measurement

Omega 24 — OPERATING INSTRUCTIONS diagram (page 16)
Fig. 7

Figure 5. 3-Wire Resistance Measurement

Page 17 of the Omega 24 user manual — 5. DATA LOGGING AND RECALL
Page 17 · click to enlarge
Page 17

WHT

Omega 24 — WHT diagram (page 17)
Fig. 8

BLK

GRN

RED

Figure 6. 4-Wire Resistance Measurement

Lead resistance errors are compensated for in 3-wire configurations. However full compensation requires equal resistance in each lead. This configuration is common with 100-ohm RTDs.

In 4-wire configurations, accuracy is unaffected by lead resistance, and resistance differences between leads.

3. Use the sensor-select key to set the instrument to the appropriate function and range.

Thermistors are identified by a momentary alpha-numeric readout of "Y-400" (signifying YSI Series-400 type thermistor).

Any of these readouts can be recalled by turning the unit off and back on.

5. DATA LOGGING AND RECALL

Operating as a meter, the unit can save and recall up to 10 readings (each with independent parameters: eg. °F/°C, TC-type and resolution) in random access memory.

Page 18 of the Omega 24 user manual — NOTES:
Page 18 · click to enlarge
Page 18

To Save Readings:

1. While in the METER OPR mode, depress the STORE key. Note that the OPR annunciator is replaced by STO.
2. Next depress one of the memory location numbers (keys 0, 1, . . ., 9). The corresponding memory annunciator turns on and the meter returns momentarily to the OPR mode, displaying thermocouple temperature.

eg. STORE 7

NOTES:

  • The enabled location annunciator indicates that the meter reading at the instant that the location number key was hit, was stored in memory.
  • When data is stored in a location already filed with data, the new data replaces the old.
  • Stored readings are retained during power-off.

To Recall Readings:

1. While in the METER OPR mode, readings stored earlier can be displayed by first depressing the RECALL key. Note that the OPR annunciator is replaced by RCL, and all memory location annunciators are blanked.
2. Next depress the keypad number corresponding to the desired location. Note that the annunciator for this location turns on, and the memory contents are displayed.
3. Several stored readings can be randomly recalled by a key sequence such as:
Omega 24 — NOTES: diagram (page 18)
Fig. 9

Figure 7.

Show 4 more labels / short text on this page
T1 Calibrator · T2 Thermometer · (output) · (input)
Page 19 of the Omega 24 user manual — NOTES:
Page 19 · click to enlarge
Page 19

eg. RECALL 3 eg. RECALL 5 eg. RECALL 2 eg. etc.

4. To return the meter to the operate mode, depress OPR after recalling the last reading. All active location annunciators will turn on again, together with the OPR annunciator.

To Erase Data Locations (Meter Mode):

1. Individual data locations can be erased by recalling the location to be cleared, followed by a CLEAR key input. Note that the annunciator for the cleared location is turned off, and the meter returns to OPR.

eg. RECALL 3 eg. CLEAR

6. CALIBRATOR MODE OPERATION (Thermocouples)

Set-Up:

1. Turn on the instrument by depressing the ON/OFF key.
2. Check that the BAT annunciator turns off after display check. If not, refer to Battery Installation/Replacement instructions.
3. When operating as a calibrator, the display indicates CALIB. Should the instrument be in the METER mode, hit the CALIB key to change to the calibrator function.
4. Select the thermocouple type that will be used.

NOTE: Space and legibility constraints on the display preclude the use of dedicated annunciators for the less common tungsten- rhenium thermocouple types (G, C, & D). Instead, these types are annunciated momentarily on the alpha-numeric display when first selected. Both T and R annunciators remain on to indicate a Tungsten-Rhenium selection. To recall the actual G, C, D selec- tion, turn the unit off, then on again for another momentary read- out.

5. Select the desired temperature scale (°F/°C key).
Page 20 of the Omega 24 user manual — CHANGE/ENTER, 2, 1, 2, 0, CHANGE/ENTER
Page 20 · click to enlarge
Page 20
6. Select the resolution desired (0.1°/1° key).
7. The unit accepts numeric inputs by first changing the keypad to its number-entry mode (NUM) by depressing the CHANGE/ENTER key once. The numeric display will flash, the OPR annunciator will turn off, and the NUM annunciator will turn on. Key in desired temperature value, most significant digit first. After the desired temperature value is keyed in, hit CHANGE//ENTER to return to OPR mode.
7. To enter a value of 212° with 0.1° resolution, the following keystrokes are used:

CHANGE/ENTER, 2, 1, 2, 0, CHANGE/ENTER

NOTES:

  • Invalid numeric entries (eg. a temperature beyond the span of the selected TC type) will cause a corresponding error code to be displayed.
  • To enter negative (ie. minus) temperatures, hit the –0 key first when in the NUM mode. The first hit on this key enters the negative sign. Subsequent hits enter zeros.
  • To correct a numeric entry while in the NUM mode, hit the CHANGE/ENTER key twice. This will return number entry to the beginning of a new number.
  • To return the calibrator output to 0° while in the OPR mode, hit the CLEAR key.
  • The unit can be used as a convenient °F to °C or °C to °F calculator.

For example, to convert 77°F to a Celsius reading, go to the CALIB OPR mode, enter the reading to be converted, and change scale.

eg. °F eg. CHANGE/ENTER eg. 77 eg. CHANGE/ENTER eg. °C (Display reads 25°C)

7. CALIBRATOR MODE OPERATION (RTD, Thermistor & Ohms)

Set-Up:

1. Connect measurement equipment under test (eg. RTD thermometer, ohmmeter, etc.) to the instrument as shown in Figures 8, 9, or 10.
Page 21 of the Omega 24 user manual — GRN
Page 21 · click to enlarge
Page 21

mA (current) source Calibrator (out) Thermometer (in)

Omega 24 — CHANGE/ENTER, 2, 1, 2, 0, CHANGE/ENTER diagram (page 21)
Fig. 10

Figure 8. 2-Wire Resistance Simulation

Omega 24 — CHANGE/ENTER, 2, 1, 2, 0, CHANGE/ENTER diagram (page 21)
Fig. 11

Figure 9. 3-Wire Resistance Simulation

Omega 24 — CHANGE/ENTER, 2, 1, 2, 0, CHANGE/ENTER diagram (page 21)
Fig. 12

GRN

RED

BLK

WHT NOTEColors indicated apply to

Model 84937 Calibration Cable.

Figure 10. 4-Wire Resistance Simulation

I out · V out

Page 22 of the Omega 24 user manual — NOTES:
Page 22 · click to enlarge
Page 22
1. The same performance tradeoffs apply to 2, 3 and 4-wire simulations as to 2, 3, and 4-wire measurements. Refer to RTD, Thermistor and Ohms Meter-Mode section above for wiring guidelines.
2. Set instrument to CALIBRATOR mode.
3. Use the sensor-select key to set the instrument to the appropriate function and range.

Thermistors are identified by a momentary alpha-numeric readout of "Y-400" (signifying YSI Series-400 type thermistor).

Any of these readouts can be recalled by turning the unit off and back on.

8. STORAGE AND RECALL OF CALIBRATOR SETTINGS

While operating as a calibrator, the unit can store and recall up to 10 output settings (each with independent parameters, eg. °F/°C, sensor type, and res- olution) in random access memory.

To Store Outputs:

1. While in the CALIB OPR mode, depress the STORE key. Note that the OPR annunciator is replaced by STO.
2. Next depress one of the memory location numbers (keys 0, 1, 2, . . ., 9). The corresponding memory annunciator turns on, and the calibrator returns to the OPR mode.

eg. STORE 2

NOTES:

  • The enabled memory annunciator shows that the output set-up has been saved.
  • When a set up is stored in a location already in use, the new set-up replaces the old.
  • Saved set-ups are retained during power-off.
  • To store a reading of 0° just hit the CLEAR while in the CALIB OPR mode. Then depress STORE and a location number.
Page 23 of the Omega 24 user manual — NOTES:
Page 23 · click to enlarge
Page 23

To Recall Outputs:

1. While in the CALIB OPR mode, set-ups saved earlier in memory can be recalled to program the calibrator output. First depress the RECALL key to recall a set-up.
1. Note that the RCL annunciator turns on, OPR turns off, and all memory locations annunciators are blanked.
2. Next hit the key with the number corresponding to the desired memory location. The corresponding annunciator turns on, and the LCD updates to show the new calibrator setting.

eg. RECALL 8

3. Several stored set-ups can be randomly recalled by a key sequence such as:

eg. RECALL 6 eg. RECALL 3 eg. RECALL 9 eg. etc.

4. To return the unit to the CALIB OPR mode after recalling the last set-up, hit the OPR key.

NOTES:

  • Any other key except ON/OFF and CLEAR will also return the OPR mode.
  • Set-ups recalled from unused locations will display and output 0°.
  • Calibrator set-ups can be recalled even when the unit is functioning in the meter mode. The same 10 storage locations are shared in both meter and calibrator modes.
  • Readings stored in memory while operating as a METER, can be recalled in the CALIB mode. In this unusual application, the unit generates calibrator outputs that simulate earlier stored meter inputs.

To Erase Set-Up Locations (Calibrator Mode):

1. Individual set-up locations can be erased by recalling the location followed by a hit to the CLEAR key.

eg. RECALL 5 eg. CLEAR

Page 24 of the Omega 24 user manual — 9. STEP FUNCTION
Page 24 · click to enlarge
Page 24

9. STEP FUNCTION

When operating in the CALIBRATOR mode, the STEP key will sequentially recall data stored in memory (up to 10 steps). At each step, the calibrator will display and output these parameters. The STEP function is non-functional in METER mode.

10. RAMP FUNCTION: CL25 and CL27

This feature, available only on Models CL25 and CL27, allows the calibrator to automatically ramp up and down a user-defined staircase. Up and down keys allow for manual ramping by the user.

To activate the RAMP function, the instrument must first be in the CALI- BRATE mode. Depress the RAMP key once to enter the ramp programming mode. The display reads "prog", with memory annunciators 1-4 flashing. A second hit to the RAMP key starts automatic ramping. A third hit turns off the RAMP function.

User definable RAMP memory locations 1, 2, 3, & 4 are dedicated to storage of RAMP parameters; step size, staircase starting point, upper limit, and lower limit respectively. These memory locations are accessible only when in RAMP “prog” mode, and are independent of the other 10 memory locations. Numeric values for RAMP parameters can be set, stored, recalled, and cleared like other calibrator settings. (Refer page 17, CALIBRATOR MODE OPERATION).

When automatic ramping is in progress, a hit to either the 7 or 8 key activates manual up-ramping or down-ramping respectively. Whenever the battery is replaced (and memory backup is not performed), the ramp parameters default to factory-set values:

LOC 1 Step Size =10°F (K TC) LOC 2 Start Point = 500°F LOC 3 High Limit = 550°F LOC 4 Low Limit = 450°F

Page 25 of the Omega 24 user manual — A. THERMOCOUPLE CALIBRATION
Page 25 · click to enlarge
Page 25

SERVICE INFORMATION

WARNING

All service information is intended for qualified electronic mainte- nance personnel only.

1. CALIBRATION PROCEDURES This is a two-part procedure. Products with thermocouple functions only, follow the Thermocouple Calibration procedure below. Products with resistance functions (RTD, thermistor, ohms) follow first the Thermocouple Calibration procedure, then the Resistance Calibration procedure.

A. THERMOCOUPLE CALIBRATION

Test Equipment Required:

1. Thermocouple simulator (OMEGA CL521, or equivalent) calibrated to ITS-90.
2. Microvolt DMM (OMEGA OM7551, or equivalent)
3. Calibration cables, per Figures 11, 12, 13.
4. Calibration cover (OMEGA 820-307-4).

Ambient Conditions:

Units should be calibrated at an ambient temperature of 23°C ±1°C, with rel- ative humidity less than 80%.

Preparation for Calibration:

11. Remove battery from unit under test (U.U.T.). Refer to Battery Installation/Replacement section in the manual.

12. Remove bottom-cover from U.U.T. Refer to Disassembly Instructions.

13. Remove calibration-jumpers J1 and J2 from printed-circuit board.

Page 26 of the Omega 24 user manual — SERVICE INFORMATION
Page 26 · click to enlarge
Page 26

14. Install calibration cover in place of bottom-cover.

15. Re-install battery.

16. Hook up test-equipment, calibration cables, and U.U.T. per Figure 10.

17. Turn on DMM and TC simulator for warmup. Allow at least 30 minutes.

18. Set the DMM to 200mV DC range.

19. Set the TC simulator to its COPPER output mode, with a reference tem- perature of 32.0°F.

10. Turn on U.U.T. and verify that "RTD" and " Ω " annunciators are flashing on the display. This verifies removal of J1 and J2 respectively.

Flashing 'Ω' indicates that error corrections for the cold-junction sensors can be stored or recalled from EEPROM.

11. Clear the EEPROM locations used to store E1 (meter cold-junction sensor error) and E2 (calibrator cold-junction sensor error) as follows:

a. Set U.U.T to CALIB, OPR, K, °F mode.

b. Key inCHANGE/0.0/ENTER/STO/1/OPR (E1 is set to 0.0°F)
c. Key inCHANGE/0.0/ENTER/STO/2/OPR (E2 is set to 0.0°F)

d. Perform RECALL 1 and RECALL 2 to verify storage of 0.0°F in both

locations.

e. Set U.U.T. to METER, OPR, K, °F.

f. Turn U.U.T. OFF, then ON to re-enable 'RTD' and 'Ω' annunciators.

Page 27 of the Omega 24 user manual — SERVICE INFORMATION
Page 27 · click to enlarge
Page 27

Calibration Adjustments:

1. Set meter-mode zero and gain adjustments per Table 1.
2. Hit OPR key (RTD annunciator turns off).
3. Set U.U.T. to CALIB, OPR, K, °F.
4. Turn U.U.T. OFF, then ON, to re-enable RTD & Ω annunciators.
5. Set calibrator-mode zero and gain adjustments per Table 2. Hit CHANGE and ENTER at the completion of each step in Table 2.
6. Hit OPR key (RTD annunciator turns off).
7. Install J1 jumper.
8. Set U.U.T. to METER, OPR, T, °F
9. Change TC simulator output to Type T, 32°F, ALLOY mode.
10. Reconfigure calibration set-up per Figure 12. Allow 2-3 minutes for thermal stabilization.
11. Calculate E1 as follows: E1 = U.U.T. Reading — 32.0.
12. Set U.U.T. to CALIB, OPR, T, °F mode.
13. Store E1; CHANGE/(E1 value)/ENTER/STO/1/OPR.
14. Go to METER mode, and verify U.U.T. reading of 32.0 ±0.1°F.
15. Reconfigure calibration setup per Figure 12. Leave simulator output at 32°F, Type T, ALLOY mode. Allow 2-3 minutes for thermal stabilization.
16. Set U.U.T. to CALIB mode, 32°F, Type T.
17. Calculate E2 as follows: E2 = DMM Reading (in µV)

–22 (For example with DMM reading of +11µV, E2 = –0.5°F)

18. Store E2; CHANGE/(E2 value)/ENTER/STO/2/OPR.
19. Install J2 jumper.
20. Re-set U.U.T. output to 32°F, DMM reading should be 0±2µV.
21. Turn OFF U.U.T.
22. Thermocouple calibration is complete. Remove calibration cover. Reinstall original back-cover, unless going on to Part B (Resistance Calibration). 25
Page 28 of the Omega 24 user manual — SERVICE INFORMATION
Page 28 · click to enlarge
Page 28
Table 1Meter-Mode Zero & Gain Adjustments

Sensor U.U.T. U.U.T. Step Select Input Adjust Reading*

1 K 0.000 mV P1 32.0 ±0.1°F 2 K 54.856 mV P11 2500.0 ±0.1°F

3 Repeat Steps 1 & 2 As Required

4 J 0.000 mV P1 32.0 ±0.1°F 5 T 0.000 mV P1 32.0 ±0.1°F 6 E 0.000 mV P1 32.0 ±0.1°F 7 N 0.000 mV P1 32.0 ±0.1°F 8 B 1.439 mV P1 1000.0 ±0.5°F 9 R 0.000 mV P1 32.0 ±0.5°F 10 S 0.000 mV P1 32.0 ±0.5°F 11 **G** 2.238 mV P1 600.0 ±0.2 °F 12 C 0.000 mV P1 32.0 ±0.2°F 13 D 0.000 mV P1 32.0 ±0.2°F

** exclusive of noise. ** on EC models, G replaced by L. Calibrate L per J.

Table 2Calibrator-Mode Zero & Gain Adjustments

Sensor U.U.T. DMM Step Select Output Adjust Reading*

1 K 32.0°F P1 0.000 mV ±1µV 2 K 2500.0°F P41 54.856 mV ±1µV

3 Repeat Steps 1 & 2 As Required

4 J 32.0°F P1 0.000 mV ±1µV 5 T 32.0°F P1 0.000 mV ±1µV 6 E 32.0°F P1 0.000 mV ±1µV 7 N 32.0°F P1 0.000 mV ±1µV 8 B 1000.0°F P1 1.439 mV ±1µV 9 R 32.0°F P1 0.000 mV ±1µV 10 S 32.0°F P1 0.000 mV ±1µV 11 **G** 600.0°F P1 2.238 mV ±1µV 12 C 32.0°F P1 0.000 mV ±1µV 13 D 32.0°F P1 0.000 mV ±1µV

** exclusive of noise. ** on EC models, G replaced by L. Calibrate L per J.

Page 29 of the Omega 24 user manual — SERVICE INFORMATION
Page 29 · click to enlarge
Page 29
Omega 24 — SERVICE INFORMATION diagram (page 29)
Fig. 13

Figure 11. Calibration Setup, Copper Mode.

Omega 24 — SERVICE INFORMATION diagram (page 29)
Fig. 14

Figure 12. Calibration Setup, Meter Mode.

Omega 24 — SERVICE INFORMATION diagram (page 29)
Fig. 15

Figure 13. Calibration Setup, Calibrator Mode.

Page 30 of the Omega 24 user manual — B. RESISTANCE CALIBRATION
Page 30 · click to enlarge
Page 30

B. RESISTANCE CALIBRATION

Test Equipment Required:

1. Resistance Decades (General Resistance RTD-100, RTD-500/1000, and RDS-54, or equivalent).
2. Current Source (Fluke 5101B, or equivalent).
3. Digital Multimeter (OMEGA OM7551, or equivalent).
4. Calibration cables per Figure 14.

Ambient Conditions:

Unit should be calibrated at an ambient temperature of 23°C±1°C, with rela- tive humidity less than 80%.

Preparation For Calibration:

1. Turn off unit, and remove bottom-cover. Refer to Disassembly Instructions. 2. Secure battery with a piece of adhesive tape.
3. Hook up test-equipment and unit per Figure 14.
4. Set current-source to 1mA, STANDBY mode.
5. Set DMM to 2V DC range.
6. Turn on unit.
7. Set unit to METER mode, RTD-100, °F.
8. Turn off unit. Remove jumper J1. Turn on unit. 'RTD' annunciator should be flashing.

will be written to memory. Avoid selecting thermocouple functions while in this mode.

functions, first hit the OPR key (disables flashing 'RTD'), and start again at Step 7 above. 28

Page 31 of the Omega 24 user manual — B. RESISTANCE CALIBRATION
Page 31 · click to enlarge
Page 31
Omega 24 — B. RESISTANCE CALIBRATION diagram (page 31)
Fig. 16

Figure 14. Resistance Calibration Setup

Page 32 of the Omega 24 user manual — (RTD-100/1000)
Page 32 · click to enlarge
Page 32

Calibration Adjustments:

1. Perform meter-mode adjustments and verifications per Table 3 (use Table 5 for RTD-Thermistor models).
2. Hit OPR key (Flashing 'RTD' annunciator turns off).
3. Install J1 jumper.
4. Set unit to CALIB mode, RTD-100, °F.
5. Switch current source to OPERATE mode.

6 Perform calibrator-mode adjustments and verifications per Table 4 (use Table 6 for RTD-Thermistor models).

7. Turn off unit. Remove tape holding battery.
8. Re-install back cover. Resistance-function calibration is complete.
Table 3Calibration of Meter-Mode Resistance Functions

(RTD-100/1000)

Sensor Unit Unit Step Select Input Adjust Verify Reading*

1 RTD-100 390.26Ω P1** 1562.0°F±0.5°F 2 RTD-100 100.00Ω √ 32.0±0.1°F 3 RTD-100 18.49Ω √ –328.0±0.2°F 4 RTD-1000 3,902.6Ω P1** 1562.0°F±0.5°F 5 RTD-1000 1000.0Ω √ 32.0+0.1°F 6 1,000Ω 900.00Ω P1 900.00Ω 7 1,000Ω 1.00Ω √ 1.00±0.10Ω 8 10,000Ω 9,000.0Ω P1 9000.0Ω 9 10,000Ω 10.0Ω √ 10.0±1.0Ω

* exclusive of noise. ** adjust as necessary to obtain steps 2, 5.

Page 33 of the Omega 24 user manual — (RTD-100/1000)
Page 33 · click to enlarge
Page 33
Table 4Calibration of Calibrator-Mode Resistance Functions

(RTD-100/1000)

Current Sensor Unit Source DMM Step Select Output Input Adjust Verify Reading*

1 RTD-100 1562.0°F 1mA P38 0.39026v 2 RTD-100 32.0°F 1mA √ 0.10000 ±0.00003v 3 RTD-100 –328.0°F 1mA √ 0.01849 ±0.00006v 4 RTD-1000 1562.0°F 100µA P51 0.39026v 5 RTD-1000 32.0°F 100µA √ 0.10000 ±0.00003v 6 1,000Ω 900.00Ω 1mA √ 0.90000 ±0.00010v 7 1,000Ω 1.00Ω 1mA √ 0.00100 ±0.00010v 8 10,000Ω 9,000.0Ω 100µA √ 0.90000 ±0.00010v 9 10,000Ω 10.0Ω 100µA √ 0.00100 ±0.00010v

* exclusive of noise.

Table 5Calibration of Meter-Mode Resistance Functions

Sensor Unit Unit Step Select Input Adjust Verify Reading*

1 RTD-100 390.26Ω P1 1562.0°F 2 RTD-100 100.00Ω √ 32.0±0.3°F 3 RTD-100 18.49Ω √ –328.0±0.5°F 4 1,000Ω 900.00Ω P1 900.00Ω 5 1,000Ω 1.00Ω √ 1.00±0.20Ω 6 100,000Ω 90,000Ω P1 90,000Ω 7 100,000Ω 100Ω √ 100±20Ω 8 Thermistor 47.0Ω P1 293.0°F 9 Thermistor 2252Ω √ 77.0±0.2°F 10 Thermistor 75,790Ω √ –40.0±0.2°F

* exclusive of noise.

(RTD-Thermistor)

Page 34 of the Omega 24 user manual — 2. CALIBRATION VERIFICATION
Page 34 · click to enlarge
Page 34
Table 6Calibration of Calibrator-Mode Resistance Functions

Current Sensor Unit Source DMM Step Select Output Input Adjust Verify Reading*

1 RTD-100 1562.0°F 1mA P38 0.39026v 2 RTD-100 32.0°F 1mA √ 0.10000 ±0.00006v 3 RTD-100 –328.0°F 1mA √ 0.01849 ±0.00012v 4 1,000Ω 900.00Ω 1mA √ 0.90000 ±0.00020v 5 1,000Ω 1.00Ω 1mA √ 0.00100 ±0.00020v 6 100,000Ω 90,000Ω 10µA P51 0.90000v 7 100,000Ω 100Ω 10µA √ 0.00100 ±0.00020v 8 Thermistor 293.0°F 1mA √ 0.04700 ±0.00020v 9 Thermistor 77.0°F 10µA √ 0.02252 ±0.00020v 10 Thermistor –40.0°F 10µA √ 0.75790 ±0.00500v

* exclusive of noise.

2. CALIBRATION VERIFICATION

Calibration is verified with the same instrument hookups as used for calibra- tion (ie. Figures 12 and 13 for thermocouple functions, Figure 14 for resis- tance functions).

While one hookup serves to verify all resistance functions (RTD, thermistor, ohms), a full check-out of thermocouple functions requires a set of calibration cables for each thermocouple type.

To simplify the verification of thermocouple calibration, software has been enhanced to reactivate the copper-mode used during calibration. This feature is available on products that use the following (and later) software revisions:

(RTD-Thermistor)

Page 35 of the Omega 24 user manual — COPPER MODE VERIFICATION PROCEDURE FOR THERMOCOUPLE FUNCTIONS
Page 35 · click to enlarge
Page 35

Model Software Revision CL23A 6.7 CL24 7.7 CL25 3.7 CL26 5.6 CL27 4.16

(Software revision is marked on a label attached to integrated-circuit U1 on the mother-board).

"Copper-mode" is a non-compensated mode of operation during which cold-junction compensation of inputs and outputs is disabled. This permits the use of copper-wire interconnections between instruments, while avoid- ing the time and expense of multiple thermocouple-cable hookups. At the conclusion of calibration verification, cold-junction compensation is re- enabled, and operation returns to "alloy-mode".

“Alloy-mode” refers to Thermocouple operations; for example, a Type-K thermocouple is made up of Nickel/Chromium and Nickel/Aluminum alloys. Referring to the OMEGA Calibrator(s) in TC operation, “Alloy-mode” means in effect “other than Copper-mode.” In order to make accurate measure- ments using a thermocouple device one must eliminate erroneous voltage signals. In “Alloy-mode”, the OMEGA calibrator automatically calculates the cold-junction compensation for the selected TC type, based on the value(s) from the ITS-90 table. This feature enables the user to make accurate tem- perature measurements without making manual calculations for error cor- rections.

COPPER MODE VERIFICATION PROCEDURE FOR THERMOCOUPLE FUNCTIONS

Equipment Required:

1. OMEGA Model to be verified.
2. Voltmeter accurate to 1µV resolution.
3. Cable assembly (Fig. 11).
1. Start your calibration with a fresh battery.
2. Press the ON function button to turn the unit on.
3. Select the calibrator mode (“CALIB” on display) by pressing the CALIB/METER function button.
Page 36 of the Omega 24 user manual — COPPER MODE VERIFICATION PROCEDURE FOR THERMOCOUPLE FUNCTIONS
Page 36 · click to enlarge
Page 36
4. Select type K thermocouple (“K” on display) by pressing the SENSOR SELECT function button until type K is reached.
5. Select the tenth degree display (with a digit appearing after the decimal point on the display) by pressing the 0.0.1 function button.
6. Select °F (“°F” on display) by pressing the °F °C function button.
7. Press the CHANGE function button. The numerical display should blink.
8. Press 9 . (“9999.9” should appear blinking on display).
9. Press the ENTER function button. The display will blink “Hi” once or twice, and then will display “0.0 °F” blinking. There will also be an “ Ω ” symbol blinking in the upper left. It will blink with the “0.0 °F” display.

10. Press the ENTER function button. The “0.0 °F” display will stop blinking, and the “Ω” will continue to blink. You are now in the copper mode. If or when you turn off the power (or remove the battery) the unit will no longer be in the copper mode and you would have to repeat the above steps to return to the copper mode. As long as you see the “Ω” blinking you are in the copper mode.

11. While in the copper mode attach the Copper Mode Calibrator Verification Cable to the connector on the top of the CL25 that is labeled T1. Attach the other end of the cable to a precision voltmeter.
12. Press the CHANGE function button. The “0.0 °F” display will blink, and “NUM” will appear on the display.
13. Press 3 2 0 buttons and then press the ENTER function button. “NUM” will disappear and “32.0 °F” will appear, not blinking. The voltmeter should read 0.000 mV.
14. Press the CHANGE function button. The “32.0 °F” display will blink, and “NUM” will appear on the display.
15. Press 2 5 0 0 0 buttons and then press ENTER function button. “NUM” will disappear and “2500.0 °F” will appear, not blinking. The voltmeter should read 54.845 mV.
16. Press the CHANGE function button. The “2500.0 °F” display will blink, and “NUM” will appear on the display.
17. Press 3 2 0 buttons and then press the ENTER function button. “NUM” will disappear and “32.0 °F” will appear, not blinking. The voltmeter should read 0.000 mV.
18. Press the SENSOR SELECT function button. The sensor type will change to type J. The “K” will disappear from the display and a “J” will appear on the display. The voltmeter should read 0.000 mV.
Page 37 of the Omega 24 user manual — COPPER MODE VERIFICATION PROCEDURE FOR THERMOCOUPLE FUNCTIONS
Page 37 · click to enlarge
Page 37
19. Press the SENSOR SELECT function button. The sensor type will change to type T. The “J” will disappear from the display and a “T” will appear on the display. The voltmeter should read 0.000 mV.
20. Press the SENSOR SELECT function button. The sensor type will change to type E. The “T” will disappear from the display and an “E” will appear on the display. The voltmeter should read 0.000 mV.
21. Press the SENSOR SELECT function button. The sensor type will change to type N. The “E” will disappear from the display and a “N” will appear on the display. The voltmeter should read 0.000 mV.
22. Press the SENSOR SELECT function button. The sensor type will change to type B. The “N” will disappear from the display and a “B” will appear on the display. “LO” will also appear on the display.
23. Press the CHANGE function button. The “32.0 °F” display will blink, and “NUM” will appear on the display.
24. Press 1 0 buttons and then press ENTER function button. “NUM” will disappear and “1000.0 °F” will appear, not blinking. The voltmeter should read 1.438 mV.
25. Press the SENSOR SELECT function button. The sensor type will change to type R. The “B” will disappear from the display and a “R” will appear on the display.
26. Press the CHANGE function button. The “1000.0 °F” display will blink, and “NUM” will appear on the display.
27. Press 3 2 0 buttons and then press ENTER function button. “NUM” will disappear and “32.0 °F” will appear, not blinking. The voltmeter should read 0.000 mV.
28. Press the SENSOR SELECT function button. The sensor type will change to type S. The “R” will disappear from the display and a “S” will appear on the display. The voltmeter should read 0.000 mV.
29. Press the SENSOR SELECT function button. The sensor type will change to type G. The “R” will disappear from the display and a “g” will appear momentarily in the center of the display. “LO” will then appear in the center of the display and both the “T” and “R” will appear along the top row of the display.
30. Press the CHANGE function button. The “32.0 °F” display will blink, and “NUM” will appear on the display.
31. Press 6 0 0 0 buttons and then press ENTER function button. “NUM” will disappear and “600.0 °F” will appear, not blinking. The voltmeter should read 2.238 mV.
Page 38 of the Omega 24 user manual — ALLOY MODE VERIFICATION PROCEDURE FOR THERMOCOUPLE FUNCTIONS
Page 38 · click to enlarge
Page 38
32. Press the SENSOR SELECT function button. The sensor type will change to type C. A “C” will appear momentarily in the center of the display. “600.0” will then appear in the center of the display and both the “T” and “R” will appear along the top row of the display.
33. Press the CHANGE function button. The “600.0 °F” display will blink, and “NUM” will appear on the display.
34. Press 3 2 0 buttons and then press ENTER function button. “NUM” will disappear and “32.0 °F” will appear, not blinking. The voltmeter should read 0.000 mV.
35. Press the SENSOR SELECT function button. The sensor type will change to type D. A “d” will appear momentarily in the center of the display. “32.0” will then appear in the center of the display and both the “T” and “R” will appear along the top row of the display. The voltmeter should read 0.000 mV.
36. Press the SENSOR SELECT function button. The sensor type will change to type K. The “T” and the “R” will disappear from the display and a “K” will appear on the display. The voltmeter should read 0.000 mV.
37. Press the OFF function button. The unit will turn off.
38. Press the ON function button. The unit will power up and return to the calibrator mode. “CALIB” will appear on the display along with the “K” enunciator, the “OPR” enunciator, and “32.0 °F” will be displayed. The blinking “ Ω ” symbol will have completely disappeared indicating that you have exited the copper mode.
39. Remove the Copper Mode Calibrator Verification Cable from both the voltmeter and the CL25.
40. The copper mode verification is now complete. Press the OFF function button to turn the unit off.

ALLOY MODE VERIFICATION PROCEDURE FOR THERMOCOUPLE FUNCTIONS

Equipment Required:

1. OMEGA Model to be verified.
2. Precision temperature simulator / voltage source.
3. Voltmeter accurate to 1µV resolution.
4. Cable assembly.
1. Start your calibration with a fresh battery.
2. Press the ON function button to turn the unit on.
Page 39 of the Omega 24 user manual — ALLOY MODE VERIFICATION PROCEDURE FOR THERMOCOUPLE FUNCTIONS
Page 39 · click to enlarge
Page 39
3. Select the calibrator mode (“CALIB” on display) by pressing the CALIB/METER function button.
4. Select type K thermocouple (“K” on display) by pressing the SENSOR SELECT function button until type K is reached.
5. Select the tenth degree display with a digit appearing after the decimal point on the display) by pressing the 0.0.1 function button.
6. Select °F (“°F” on display) by pressing the °F °C function button.

7. Attach the Alloy Mode Calibrator Verification Cables per Fig. 12 & 13. If you are using a Type K Thermocouple Temperature Simulator ITS-90 you will use the regular set up. If you are using a Precision Voltage Source use the alternate set up. When performing the following proce- dure using a Precision Voltage Source with the alternate set up, any- where a temperature is transmitted (output) by the Thermocouple Temperature Simulator, you should adjust the Voltage Source to the equivalent voltage as listed by a Type K N.I.S.T. ITS-90 Thermocouple Reference Table. (Example: 32.0 °F would equal 0.000 mV)

8. Press the CHANGE function button and the numerical display will blink and “NUM” will appear on the display.
9. Press 3 2 0 buttons and then press the ENTER function button. “NUM” will disappear and “32.0 °F” will appear, not blinking.
10. Adjust the Thermocouple Temperature Simulator to 32.0 °F.
0. Note: The reading on the voltmeter should read null. 0.000 mV. Adjusting the CL25 to any temperature and adjusting the Temperature Simulator to the same temperature should result in a null, or 0.000 mV on the voltmeter.
11. Remove the Alloy Mode Calibrator Verification Cable from all equipment.
12. The Alloy Mode verification is now complete. Press the OFF function button to turn the unit off.
Page 40 of the Omega 24 user manual — 4. MECHANICAL PARTS DIAGRAM
Page 40 · click to enlarge
Page 40

4. MECHANICAL PARTS DIAGRAM

Omega 24 — 4. MECHANICAL PARTS DIAGRAM diagram (page 40)
Fig. 17
Page 41 of the Omega 24 user manual — CALIBRATION PROCEDURES Test Equipment Required:
Page 41 · click to enlarge
Page 41

CALIBRATION PROCEDURES Test Equipment Required:

1. Thermocouple simulator (OMEGA CL25, or equivalent).
2. Microvolt null-meter (Keithley 155, or equivalent)
3. Calibration cover (OMEGA 820-307-4).
4. Calibration cables, T.C. types K, J, T & E (as required) per Figure1 below.
5. Low-thermal (eg. copper) shorting-link for null-meter.

Ambient Conditions:

Units should be calibrated at an ambient temperature of 23°C ±1°C, with rel- ative humidity less than 80%. Test equipment and the unit-under-test (U.U.T.) should have 1 hour to reach thermal equilibrium with ambient-tem- perature.

Calibration Adjustments:

11. Remove battery from unit under test (U.U.T.). Refer to Battery Installation/Replacement instructions in the manual.

12. Remove bottom-cover from U.U.T. Refer to Disassembly Instructions.

13. Remove calibration-jumper J1 from printed-circuit board.

14. Install calibration cover in place of bottom-cover.

15. Re-install battery.

16. Connect U.U.T. to test equipment per Figure 2 using the K-type calibra- tion cable.

17. Turn on U.U.T. and allow 2-3 minutes for thermal re-stabilization of the TC mini-plug and input-jack.

annunciator is on, indicating that new constants will be stored in EEPROM.

Page 42 of the Omega 24 user manual — CALIBRATION PROCEDURES Test Equipment Required:
Page 42 · click to enlarge
Page 42

abled. This is to speed up U.U.T. response to calibra- tion adjustments. In this mode ±0.1°F of reading noise is normal.

sequence that follows. This will ensure that zero adjust- ments for each TC type will be stored in the appropriate memory locations.

18. Apply 32°F (Type K) to U.U.T. Adjust potentiometer P1 for a reading of 32.0°F (exclusive of noise).

19. Apply 2500°F (Type K) to U.U.T. Adjust potentiometer P11 for a reading of 2500.0°F.

10. Hit OPR key to store K-meter zero-adjustment in EEPROM. Note that “FUNC” annuciator turns off.
11. Select Type J thermocouple (KJT key on CL23A, SENSOR SELECT key on CL24).
12. Turn off U.U.T.
13. Replace K calibration-cable with J calibration-cable.
14. Turn on U.U.T. Note that “FUNC” annunciator is on again. Allow 2-3 minutes for thermal restabilization.
15. Apply 32°F (Type J). Adjust P1 for a reading of 32.0°F (exclusive of noise).
16. Hit OPR key to store J-meter zero-adjustment.
17. Select Type T thermocouple.
18. Turn off U.U.T.
19. Replace J calibration-cable with T calibration-cable.
20. Turn on U.U.T. Allow 2-3 minutes for thermal stabilization.
21. Apply 32°F (Type T). Adjust P1 for a reading of 32.0°F (exclusive of noise).
22. Hit OPR key to store T-meter zero-adjustment.
Page 43 of the Omega 24 user manual — CALIBRATION PROCEDURES Test Equipment Required:
Page 43 · click to enlarge
Page 43
23. Select Type-E thermocouple.
24. Turn off U.U.T.
25. Replace T calibration-cable with E calibration-cable.
26. Turn on U.U.T. Allow 2-3 minutes for thermal stabilization.
27. Apply 32°F (Type E). Adjust P1 for a reading of 32.0°F (exclusive of noise).
28. Hit OPR key to store E-meter zero-adjustment.
29. Select Type-K thermocouple.
30. Select CALIB mode.
31. Turn off U.U.T.
32. Connect U.U.T. to test-equipment per Figure 3 using the K calibrationcable.
33. Turn-on U.U.T. and allow 2-3 minutes, for thermal stabilization
34. Set both the U.U.T. and TC-simulator to output 32.0°F (Type K). Adjust P2 for a null-meter reading of 0±3µV.
35. Set both the U.U.T. and T-simulator to output 2400°F (Type K). Adjust P41 for a null-meter reading of 0±3µV.
36. Hit OPR key to store K-calib zero adjustment.
37. Select J thermocouple.
38. Turn off U.U.T.
39. Replace K calibration-cable with J calibration-cable.
40. Turn on U.U.T. Allow 2-3 minutes for thermal stabilization.
41. Set both the U.U.T. and TC-simulator to output 32.0°F (Type J). Adjust P1 for a null-meter reading of 0±3µV.
42. Hit OPR Key to store J-calib zero adjustment.
43. Select Type-T thermocouple.
44. Turn off U.U.T.
45. Replace J calibration-cable with T calibration-cable.
46. Turn on U.U.T. Allow 2-3 minutes for thermal stabilization.
Page 44 of the Omega 24 user manual — CALIBRATION PROCEDURES Test Equipment Required:
Page 44 · click to enlarge
Page 44
47. Set both the U.U.T. and TC-simulator to output 32.0°F (Type T). Adjust P1 for a null-meter reading of 0±3µV.
48. Hit OPR key to store T-calib zero adjustment.
49. Select Type-E thermocouple.
50. Turn off U.U.T.
51. Replace T calibration-cable with E calibration-cable.
52. Turn on U.U.T. Allow 2-3 minutes for thermal stabilization.
53. Set both the U.U.T. and thermocouple-simulator to 32.0°F (Type E). Adjust P1 for a null-meter reading of 0±3µV.
54. Hit OPR Key to store E-calib zero adjustment
55. Turn off U.U.T.
56. Remove calibration cover. Be careful not to disturb battery connections.
57. Re-install J1 jumper.
58. Re-install original back cover. Calibration is complete.
Page 45 of the Omega 24 user manual — NOTES
Page 45 · click to enlarge
Page 45

NOTES

Page 46 of the Omega 24 user manual — NOTES
Page 46 · click to enlarge
Page 46

NOTES

Page 47 of the Omega 24 user manual — MADE IN
Page 47 · click to enlarge
Page 47

MADE IN

USA

WARRANTY/DISCLAIMER

OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 37 months from date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal three (3) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.

If the unit should malfunction, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of being damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not warranted, including but not limited to contact points, fuses, and triacs.

OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MER- CHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMI- TATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the compo- nent upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.

CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/DISCLAIMER language, and additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.

RETURN REQUESTS / INQUIRIES Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence. The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.

FOR WARRANTY RETURNS, please have the fol- lowing information available BEFORE contacting OMEGA: 1. P.O. number under which the product was PURCHASED, 2. Model and serial number of the product under warranty, and 3. Repair instructions and/or specific problems relative to the product.

FOR NON-WARRANTY REPAIRS, consult OMEGA for current repair charges. Have the fol- lowing information available BEFORE contacting OMEGA: 1. P.O. number to cover the COST of the repair, 2. Model and serial number of product, and 3. Repair instructions and/or specific problems relative to the product.

OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering. OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 1996 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without prior written consent of OMEGA ENGINEERING, INC.

Page 48 of the Omega 24 user manual — M1554/ 1105
Page 48 · click to enlarge
Page 48

Where Do I Find Everything I Need for

Process Measurement and Control?

OMEGA…Of Course!

TEMPERATURE M U Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies MU Wire: Thermocouple, RTD & Thermistor MU Calibrators & Ice Point References MU Recorders, Controllers & Process Monitors MU Infrared Pyrometers PRESSURE, STRAIN AND FORCE MU Transducers & Strain Gauges MU Load Cells & Pressure Gauges MU Displacement Transducers MU Instrumentation & Accessories FLOW/LEVEL MU Rotameters, Gas Mass Flowmeters & Flow Computers MU Air Velocity Indicators MU Turbine/Paddlewheel Systems MU Totalizers & Batch Controllers pH/CONDUCTIVITY M U pH Electrodes, Testers & Accessories MU Benchtop/Laboratory Meters MU Controllers, Calibrators, Simulators & Pumps MU Industrial pH & Conductivity Equipment DATA ACQUISITION MU Data Acquisition & Engineering Software MU Communications-Based Acquisition Systems MU Plug-in Cards for Apple, IBM & Compatibles MU Datalogging Systems MU Recorders, Printers & Plotters HEATERS MU Heating Cable MU Cartridge & Strip Heaters MU Immersion & Band Heaters MU Flexible Heaters MU Laboratory Heaters ENVIRONMENTAL MONITORING AND CONTROL MU Metering & Control Instrumentation MU Refractometers MU Pumps & Tubing MU Air, Soil & Water Monitors MU Industrial Water & Wastewater Treatment MU pH, Conductivity & Dissolved Oxygen Instruments

M1554/ 1105

Ask about the Omega 24

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Omega 24

Be the first to ask — other owners or our team will answer.

Original document

Omega 24 manual PDF

48 pages·666 KB·Thermometer·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…