USER GUIDE AND SPECIFICATIONS NI USB-6008/6009 This guide describes how to use the National Instruments USB-6008/6009 data acquisition (D A Q) de vices and lists specif ications.
NI USB-6008/6 009 User Guid e and Specifications 2 ni.com Figure 1. NI USB-6008/6009 Figure 2. NI USB-6008/6009 Back View Table 2. Differences Between the NI USB-6008 and NI USB-6009 Featur e NI USB-6.
© National Instruments Corporation 3 NI US B-6008/6009 User G u ide and Specifications Safety Guidelines Caution Operate the hardware only as describe d in these operating instructions. The following section contains important safety information that you must follow when installing and using the NI USB-6008/6009.
NI USB-6008/6 009 User Guid e and Specifications 4 ni.com Operate the device at or belo w the Measurement Cate gory I 1 . Measurement circuits are subjected to working v oltages 2 and transient stresses (ov ervoltage) from the circuit to which they are connected during measurement or test.
© National Instruments Corporation 5 NI US B-6008/6009 User G u ide and Specifications Software Software support for t he NI USB-6008/6009 fo r Windows 2000/XP/Vista is provided by NI-DAQmx . The NI-D A Qmx CD contains example progra ms that you can use to get started programming with the NI USB-6008/ 6009.
NI USB-6008/6 009 User Guid e and Specifications 6 ni.com Hardware The following block diagram show s key functional components of the NI USB-6008/6009. Figure 3. Device Bl ock Diagram P1.<0..3> P0.<0..7> Digital I/O T erminal Block AI <0.
© National Instruments Corporation 7 NI US B-6008/6009 User G u ide and Specifications Setting Up Hardware Complete the f ollowing step s to set up the ha rdware: 1. Install combicon scr e w terminal blocks by inserting them into the combicon jacks. 2.
NI USB-6008/6 009 User Guid e and Specifications 8 ni.com Figure 5. Signal Label Application Diagram Note Once you label the screw terminal blocks, you must only insert them into the matching combicon jack, as indi cated by the ov erlay label on the NI USB-6008/6009 device.
© National Instruments Corporation 9 NI US B-6008/6009 User G u ide and Specifications I/O Connector The NI USB-6008/6009 ship s with one detachable screw term inal block for analog signals and one detachable screw terminal block for digit al signals.
NI USB-6008/6 009 User Guid e and Specifications 10 ni.com Table 4. Digital T erminal Assignments Module T erminal Signal 17 P0.0 18 P0.1 19 P0.2 20 P0.
© National Instruments Corporation 11 NI US B-6008/600 9 User Gu ide and Specifications Signal Descriptions Table 5 describes the signals available on the I/O connectors.
NI USB-6008/6 009 User Guid e and Specifications 12 ni.com LED The NI USB-6008/6009 device has a green LED next to the USB connector. When the device is connect ed to a USB port, the LED blinks steadily to indicate that the device is initialized and is receiving power from the connection.
© National Instruments Corporation 13 NI US B-6008/600 9 User Gu ide and Specifications PGA The progammable-gain amplifi er provides input gains of 1, 2, 4, 5, 8, 10, 16, or 20 when configured for differential measur ements and gain of 1 when configured for single-en ded measurements.
NI USB-6008/6 009 User Guid e and Specifications 14 ni.com The dif ferential input mode can measure ± 20 V signals in the ± 20 V range. Ho wev er , the maximum voltage on any one pin is ± 10 V with respect to GND. For e xample, if AI 1 is + 10 V and AI 5 is –10 V , then th e measurement returned from the de vice is + 20 V .
© National Instruments Corporation 15 NI US B-6008/600 9 User Gu ide and Specifications Connecting Reference Singl e-Ended V oltage Signals To connect reference single-ende d volt age signals (RSE) to the NI USB-6008/6009, connect the positive voltage signal to the desired AI terminal, and the ground signal to a GND termin al.
NI USB-6008/6 009 User Guid e and Specifications 16 ni.com Analog Output The NI USB-6008/6009 has two independent AO chann els that can generate outputs from 0 –5 V. All updates of AO lines are software-timed. Analog Output Circuitry Figure 11 illustrates the analog outpu t circuitry for the NI USB-6008/6009.
© National Instruments Corporation 17 NI US B-6008/600 9 User Gu ide and Specifications Minimizing Glitches on the Output Signal When you use a DAC to generate a wa veform, y ou may observe g litches in the output signal. These glitches are no rmal; when a DAQ switches from one voltage to an other, it produces g litches due to released charges.
NI USB-6008/6 009 User Guid e and Specifications 18 ni.com Caution Exceeding the maximum input voltage ratin gs or maximum output ratings, which are listed in the Specifications section, can damage the D A Q de vice and the computer . National Instruments is not liable for any damage resulting from such signal connections.
© National Instruments Corporation 19 NI US B-6008/600 9 User Gu ide and Specifications 6. Select a static resistor v alue for your pull-up resistor th at is greater than or equal to the ideal resistance. 7. Re-connect the load circu it and the pull-up resistor .
NI USB-6008/6 009 User Guid e and Specifications 20 ni.com Reference and Power Sources The NI USB-6008/6009 creates an external reference and suppl ies a power source.
© National Instruments Corporation 21 NI US B-6008/600 9 User Gu ide and Specifications Timing accuracy ........ .............. ............... 1 00 ppm of actual sample rate Input range Single-ended .............. ................. .... ± 10 V Differential .
NI USB-6008/6 009 User Guid e and Specifications 22 ni.com Analog Output Analog outputs ...................... ................. .2 Output resolution .................. ..................12 bits Maximum update rate .................... .........150 Hz, software-timed Output range .
© National Instruments Corporation 23 NI US B-6008/600 9 User Gu ide and Specifications Compatibility ... ................. .................... . TTL, LVTTL, CMOS Absolute maximum voltag e range ......... – 0.5 to 5.8 V with respect to GND Pull-up resistor .
NI USB-6008/6 009 User Guid e and Specifications 24 ni.com Input high voltage ............................. ......2.0 V Input low voltage ..................... ...............0.8 V Bus Interface USB specification ............ ...................... .USB 2.
© National Instruments Corporation 25 NI US B-6008/600 9 User Gu ide and Specifications Safety Standards The NI USB-6008/6009 is designed to meet th e requirements of the following standards of saf e.
NI USB-6008/6 009 User Guid e and Specifications 26 ni.com Storage temperature (IEC 60068-2-1 and IEC 60068-2-2) ...... – 40 to 85 °C Storage humidity (IEC 60068-2-56) ........ .............. ............5 to 90% RH, non condensing Pollution Degree (IEC 60664) .
National Instruments, NI, ni.com, and LabVIEW are tradem arks of National Instruments Corporation. Refer to t he Terms of Use section on ni.com/legal for more informat ion about National Instruments trademarks. Other product and compan y names mentioned herein are trademarks or trade names of their respective companie s.
An important point after buying a device National Instruments 6008 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought National Instruments 6008 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data National Instruments 6008 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, National Instruments 6008 you will learn all the available features of the product, as well as information on its operation. The information that you get National Instruments 6008 will certainly help you make a decision on the purchase.
If you already are a holder of National Instruments 6008, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime National Instruments 6008.
However, one of the most important roles played by the user manual is to help in solving problems with National Instruments 6008. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device National Instruments 6008 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center