
7532 Radar Tank Gauge Smart transmitter for continuous and non-contact precision level measurement. For custody transfer and inventory-control applications with NMI and PTB approval.

Varec, Inc. 5834 Peachtree Corners East, Norcross (Atlanta), GA 30092 USA Tel: +1 (770) 447-9202 Fax: +1 (770) 662-8939


- unknown - 10
- ok - too small - too big - unknown - manual
- envel. curve - incl. FAC - incl. cust. map
555 = History Reset (333 = reset customer parameters)
An overview of all device functions can be found on page 96.
A description of the device functions for the 7532 RTG can also be found on the enclosed CD-ROM.
Radar Tank Gauge


Safety Instructions Radar Tank Gauge
1.1 Designated use
The Varec Model 7532 Radar Tank Gauge (RTG) is a compact radar level transmitter for the continuous, contactless measurement of liquids, pastes and sludge in stilling wells. The device can also be freely mounted outside closed metal vessels because of its operating frequency of about 6 GHz and a maximum radiated pulsed energy of 1mW (average power output 1 µW). Operation is completely harmless to humans and animals.
The 7532 RTG has been designed to operate safely in accordance with current technical, safety and EU standards. If installed incorrectly or used for applications for which it is not intended, however, it is possible that application-related dangers may arise, e.g. product overflow due to incorrect installation or calibration. For this reason, the instrument must be installed, connected, operated and maintained according to the instructions in this manual: personnel must be authorised and suitably qualified. The manual must have been read and understood, and the instructions followed. Modifications and repairs to the device are permissible only when they are expressly approved in the manual.
Hazardous areas Measuring systems for use in hazardous environments are accompanied by separate “Ex documentation”, which is an integral part of this Operating Manual. Strict compliance with the installation instructions and ratings as stated in this supplementary documentation is mandatory.
FCC-approval This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.

The following procedures must be carried out before a transmitter is sent to Varec for repair:
7532 Safety Instructions

Safety Instructions Radar Tank Gauge
In order to highlight safety-relevant or alternative operating procedures in the manual, the following conventions have been used, each indicated by a corresponding symbol in the margin.

2.1 Device designation
The following technical data are given on the instrument nameplate:


Figure 2: Information on the NMi type plate for custody transfer applications of the 7500 RTG (example)

Figure 3: Information on the PTB type plate for custody transfer applications of the 7500 RTG (example)
7532 Identification

Identification Radar Tank Gauge
Ordering structure 7532 RTG
Flange Dia/Pressure Standard Material AVJ 6"/150 lbs / RF ANSI B16.5 316/316L 11.4 kg A3J 8"/150 lbs / RF ANSI B16.5 316/316L 19.6 kg A5J 10"/150 lbs / RF ANSI B16.5 316/316L 28.8 kg AWJ 6"/300 lbs / RF ANSI B16.5 316/316L 20.9 kg AXJ 8"/300 lbs / RF ANSI B16.5 316/316L 34.3 kg A7J 12"/150 lbs / RF ANSI B16.5 316/316L 43.2 kg CWJ DN150 PN10 / 16 C EN 1092,1, B11) 316L 10.6 kg
CXJ DN200 PN16 C EN 1092,1, B11) 316L 16.5 kg
CZJ DN200 PN25 C EN 1092,1, B11) 316L 22.7 kg
C1J DN150 PN25 C EN 1092,1, B11) 316L 14.7 kg
C6J DN250 PN16 C EN 1092,1, B11) 316L 25.6 kg
C8J DN300 PN16 C EN 1092,1, B11) 316L 36.1 kg
XXJ with flange hub 316L XVU Varec UNI-Flange 6"/ DN150/150A, max. 14.5LBS/PN1/1K, compatible with: - 6" 150LBS - DN150 PN16 - 10K 150A
304/1.4301 3.5 kg
X3U Varec UNI-Flange 8"/ DN200/200A, max. 14.5LBS/PN1/1K, compatible with: - 8" 150LBS - DN200 PN16 - 10K 200A
304/1.4301 5.2 kg
X5U Varec UNI-Flange 10"/ DN250/250A, max. 14.5LBS/PN1/1K, compatible with: - 10" 150LBS - DN250 PN16 - 10K 250A
304/1.4301 7.5 kg
X7U Varec UNI-Flange 12"/ DN300/300A, max. 14.5LBS/PN1/1K, compatible with: - 12" 150LBS - DN300 PN16 - 10K 300A
304/1.4301 10.8 kg
30 Process connection

40 Output and operation A 4…20 mA HART with VU 331, 4-line alphanumeric dis- play Y Special version
50 Housing C Aluminium T12-housing with separate connection compartment, coated, IP65 Y Special version
60 Gland / Entry 2 M20x1.5 cable gland 4 ½ NPT cable entry 9 Special version
70 Custody transfer approvals A NMi, PTB type and test rig approval, weights&measures approved (< 1 mm) R Inventory Control Version, not weights&measures approved (3 mm) Y Special approval for custody transfer
80 Additional options A Without additional options Y Special version
7532 Identification · 7532- Complete product designation

Identification Radar Tank Gauge
The scope of delivery consists of:
Accompanying documentation:
CE mark, declaration of conformity The instrument is designed to meet state-of-the-art safety requirements, has been tested and left the factory in a condition in which it is safe to operate. The instrument complies with the applicable standards and regulations in accordance with EN 61010 “Protection Measures for Electrical Equipment for Measurement, Control, Regulation and Laboratory Procedures”. The instrument described in this manual thus complies with the statutory requirements of the EG directives. Varec confirms the successful testing of the instrument by affixing to it the CE mark.
KALREZ ®, VITON ®, TEFLON ® Registered trademark of the company E.I. Du Pont de Nemours & Co., Wilmington, USA
TRI-CLAMP ® Registered trademark of the company Ladish & Co., Inc., Kenosha, USA
HART ® Registered trademark of HART Communication Foundation, Austin, USA
ToF ® Registered trademark of the company Endress+Hauser GmbH+Co. KG, Maulburg, Ger- many
Registered trademark of the company Endress+Hauser GmbH+Co. KG, Maulburg, Ger- many
Registered trademark of the company Endress+Hauser GmbH+Co. KG, Maulburg, Ger- many
PulseMaster ® · PhaseMaster ®

3.1 Quick installation guide

The housing can be turned in order to simplify access to the display and the terminal compartment
The performance of the 7532 RTG planar antenna is not dependent on the alignment or geometry of standard stilling wells. No special alignment is required. However, make sure that the planar antenna is installed vertically relative to the stilling well axis.
7532 Mounting

Mounting Radar Tank Gauge
Check the packing and contents for any signs of damage. Check the shipment, make sure nothing is missing and that the scope of supply matches your order.
Pack the measuring instrument so that is protected against impacts for storage and transport. The original packing material provides the optimum protection for this. The permissible storage temperature is -40 °C…+80 °C.


detail A flange hub for the connection to flanges provided by the customer
O-Ring 85.3 x 3.53, included (same material as sensor seal)
Antenna version DN150 / 6" DN200 / 8" DN250 / 10" DN300 / 12"
L [mm] 93 337 490 517
Ø D [mm] no horn 190 240 290
7532 Mounting

Mounting Radar Tank Gauge
BD
max. level
B
recommended additional settings Blocking
BD [m / ft] SD [m / ft] A [mm / inch] B [m / ft] C [mm / inch]
7532 RTG (planar) 1 / 3.28 0.5 / 1.6 1000 / 40 0.5 / 1.64 150…300 / 6…12
Behaviour if measuring range is exceeded The behaviour in case of the measuring range being exceeded can be freely set: the default setting is a current of 22 mA and the generation of a digital warning (E681).

Measuring range The usable measuring range depends on the size of the antenna, the reflectivity of the medium, the mounting location, and eventual interference reflections. The following tables describe the groups of media as well as the achievable measuring range as a function of application and media group. If the dielectric constant of a medium is unknown, it is recommended to assume media group B to ensure a reliable measurement.
A 1.4 … 1.9 non-conducting liquids, e.g. liquefied gas 1)
B 1.9 … 4 non-conducting liquids, e.g. benzene, oil, toluene, …
C 4 … 10 e.g. concentrated acids, organic solvents, esters, aniline, alcohol, acetone, …
D > 10 conducting liquids, e.g. aqueous solutions, dilute acids and alkalis
1)Treat Ammonia NH3 as a medium of group A, e.g. always use a stilling well.
Measuring range depending on product class for 7532 RTG:
Product class Stilling well/ Bypass
A DK(εr)=1,4…1,9 38 m/124 ft
B DK(εr)=1,9…4 38 m/124 ft
C DK(εr)=4…10 38 m/124 ft
D DK(εr)>10 38 m/124 ft
max. measuring range with custody transfer approvals
7532 Mounting · Measuring range · 20 m/65 ft

Mounting Radar Tank Gauge
Blocking distance The blocking distance (= BD) is the minimum distance from the reference point of the measurement (mounting flange) to the medium surface at maximum level.
Max. Level
Stilling well / Bypass
from Flange 1 m/40"

In addition to the tool needed for flange mounting, you will require the following tool:
Optimum mounting position for 7532 RTG
Standard installation When mounting in the stilling well, please observe engineering tips on page 14 and the following points:

Recommendations for the stilling well
Stilling wells with pipe expansion
7532 Mounting

Mounting Radar Tank Gauge
Examples for the construction of stilling wells


After mounting, the housing can be turned 350° in order to simplify access to the display and the terminal compartment. Proceed as follows to turn the housing to the required position:

7532 Mounting

Mounting Radar Tank Gauge
Installation tips Varec UNI flanges are designed for non-pressurized operation respectively max. 1 bar absolute pressure. The number of bolts has sometimes been reduced. The bolt-holes have been enlarged for adaption of dimensions, therefore, the flange needs to be properly aligned to the counterflange before the bolts are tightened.

Version Compatible with D [mm] K [mm] Type plate no.
482 410 (for DIN), 431,8 (for ANSI), 400 (for

Preparation for the installation of the Varec UNI flange

7532 Mounting

Mounting Radar Tank Gauge
mount Varec UNI flange
Varec UNI flange (max. 1 bar)
detail A flange hub for the connection to flanges provided by the customer
Ø 62.5
O-Ring 85.3 x 3.53, included (same material as sensor seal)
pull off housing · L00-FMR532xx-00-00-06-en-003 · flange hub

Installation tips For control and cleaning purposes as well as for hand dipping (tape), a sample hatch gauging is recommended. The sensor head can be easily checked in the area of the opening. Manual gauging with gauge rod or tape is possible without removal of the transmitter. The lower edge of the opening is the reference for the gauging. The construction is only suitable for non-pressurized operation.

Flange DN150 DN200 DN250/300 Flange ANSI 6" ANSI 8" ANSI 10"
PN [lbs] 2) 16 16 16 PN [lbs] 2) 150 150 150
A [mm] 110 140 170 A [mm] 110 140 170
L [mm} — 300 450 L [mm] — 300 450
2)Only dimensions adapted to standard. Designed for non-pressurized operation only, therefore thickness of flange can be reduced (e.g. 8mm).
7532 Mounting

Mounting Radar Tank Gauge
Installation tips The Pivoting element serves to swivel the 7500 RTG from the measuring position, e.g. to clean the antenna or dip the tank. The pivoting element is not part of the standard offering from Varec, special offers available under ref. number MVTFM0422.
After the measuring instrument has been installed, perform the following checks:
L00-FMR53xxx-17-00-00-en-013

4.1 Quick wiring guide
When grounding conductive screens, the corresponding directives EN 60079-14 and EN 1127-1 must be observed. Recommendation for safe grounding of conductive screens:
7532 Wiring · Wiring

Wiring Radar Tank Gauge
Before connection please note the following:
The power supply must be identical to the data on the nameplate (1).
Switch off power supply before connecting up the device.
Connect Equipotential bonding to transmitter ground terminal before connecting up the device.
Tighten the locking screw: It forms the connection between the antenna and the housing ground potential.
When you use the measuring system in hazardous areas, make sure you comply with national standards and the specifications in the safety instructions (XA’s). Make sure you use the specified cable gland.
Connect up the 7500 RTG as follows:
Before unscrew housing cover (2) at seperate connection room turn off the power supply!
Insert cable through gland (3). Use screened, twisted wire pair.
Only ground screening of the line (5) on sensor side.
Make connection (see pin assignment).
Tighten cable gland (4).
Screw off housing cover (2).
Switch on power supply.
1 2 3 4
1 3 2 4 5
A 7500 RTG situated in a hazardous area is connected as a single device to a power supply unit and transmitter situated outside of the hazardous area. In this case, it is recommended that the screen be connected directly to the 7500 RTG at the housing's earth, whereby the 7500 RTG and the power supply unit are connected to the same potential matching line (PML).
75xx_wiring.eps

Wiring with Tank Side Monitor 4590
Before connection please note the following:
The power supply must be identical to the data on the nameplate (1).
Switch off power supply before connecting up the device.
Connect Equipotential bonding to transmitter ground terminal before connecting up the device.
Tighten the locking screw: It forms the connection between the antenna and the housing ground potential.
When you use the measuring system in hazardous areas, make sure you comply with national standards and the specifications in the safety instructions (XA’s). Make sure you use the specified cable gland.
Connect up the 7500 RTG as follows:
Before unscrew housing cover (2) at seperate connection room turn off the power supply!
Insert cable through gland (3). Use screened, twisted wire pair.
Make connection (see pin assignment).
Tighten cable gland (4).
Screw off housing cover (2).
Switch on power supply.
intrinsicaly safe terminal board
NRF 590 i.s. terminal board
RTD
S+
SD-
OPT1
grounding not on 7500 RTG grounding single sided on Tank Side Monitor 4590 TSM
OPT3
i.s. module wiring
H
1 2 3 4
Internally interconnected as one HART fieldbus loop
1 2 3 4 5
H
P
1 3 2 4 5
For Micropilot S-series only!
only for 7500 RTG
75xx_wiring_w_TSM.
The 7500 RTG is - possibly in combination with other devices - connected to a Tank Side Monitor in a hazardous area. In this case, it is recommended that you ground the cable screen centrally at the Tank Side Monitor and connect all devices to the same potential matching line (PML). If, for functional reasons, a capacitive coupling is required between local earth and screen (multiple grounding), ceramic condensers with a dielectric strength of min. 1500 Veff must be used, whereby the total capacitance of 10 nF must not be exceeded. Notes on grounding interconnected intrinsically safe devices are provided by the FISCO model

Wiring Radar Tank Gauge
Terminal compartment The housing comes with a separate terminal compartment.
Load HART Minimum load for Hart communication: 250 Ω
Supply voltage Direct current voltage: 16…36 VDC .
Communication Terminal voltage minimal maximal
Power supply Standard U (20 mA) = 16 V 36 V
Ex U (20 mA) = 16 V 30 V
U (4 mA) = 11,5 V 30 V
U (20 mA) = 11,5 V 30 V
Power consumption Max. 330 mW at 16 V, max. 500 mW at 24 V, max. 600 mW at 30 V.
Current consumption Max. 21 mA (50 mA inrush current).
Power supply For stand alone operation recommended via e.g. Varec RN 221N.
mm accuracy For measurements with mm accuracy the measured variable must be transmitted using HART protocol to ensure the necessary resolution.
L00-FMR53xxx-04-00-00-en-001 · Signal · Ex

See page 27.


7532 Wiring

Wiring Radar Tank Gauge
Connect the Equipotential bonding to the external ground terminal of the transmitter.
–Connect the external overvoltage protector and the 7500 RTG transmitter to the local
potential matching system. –Potentials shall be equalised both inside and outside the explosion hazardous area. –The cable connecting the overvoltage protector and the 7500 RTG transmitter shall
not exceed 1 m in length; –The cable shall be protected e.g. routed in an armoured hose.
After wiring the measuring instrument, perform the following checks:

5.1 Quick operation guide
7532 Operation

Operation Radar Tank Gauge
F
O O O F F
S S
S
Selection and configuration in Operation menu:
1.) Change from Measured Value Display to by pressing
2.) Press or to select the required (e.g.. "basic setup (00)") and confirm by pressing (e.g. "tank shape (002)") is selected.
The active selection is marked by a in front of the menu text.
S
3.) Activate Edit mode with or .
Selection menus:
a) Select the required in selected (e.g. "tank shape (002)") with or . b) confirms selection appears in front of the selected parameter c) confirms the edited value system quits Edit mode d) + (= ) interrupts selection system quits Edit mode
F F O S X
Typing in numerals and text:
a) Press or to edit the first character of the (e.g. "empty calibr. (005)") b) positions the cursor at the next character (a) until you have completed your input c) if a symbol appears at the cursor, press to accept the value entered system quits Edit mode d) + (= ) interrupts the input,
F
O X F O S X O S O X
S
system quits Edit mode
Press + (= ) twice return to

The operating menu is made up of two levels:
If, for example, the application of the instrument is to be changed, carry out the follow- ing procedure:
1.Select the “basic setup (00)” function group. 2.Select the "tank shape (002)" function (where the existing tank shape is selected).
For simple orientation within the function menus (see page 91), for each function a position is shown on the display.

The first two digits identify the function group:
The third digit numbers the individual functions within the function group:
Here after the position is always given in brackets (e.g. "tank shape" (002)) after the described function.

Operation Radar Tank Gauge

Liquid crystal display (LCD): Four lines with 20 characters each. Display contrast adjustable through key combination.


The following table describes the symbols that appear on the liquid crystal display:
ALARM_SYMBOL This alarm symbol appears when the instrument is in an alarm state. If the symbol flashes, this indicates a warning.
LOCK_SYMBOL This lock symbol appears when the instrument is locked,i.e. if no input is possible.
COM_SYMBOL This communication symbol appears when a data transmission via e.g. HART, PFOFIBUS-PA or Foundation Fieldbus is in progress.
Calibration to regulatory standards disturbed If the instrument is not locked or it cannot guarantee the calibration to regulatory standards, the situation will be indicated on the display via the symbol.
Light emitting diodes (LEDs): There is a green and a red LED besides the Liquid Crystal Display.
red LED continuously on Alarm
red LED flashes Warning
red LED off No alarm
green LED continuously on Operation
Green LED flashes Communication with external device
7532 Operation · Symbols Meaning · LED Meaning

Operation Radar Tank Gauge
The operating elements are located inside the housing and are accessible for operation by opening the lid of the housing.
Function of the keys
O or V Navigate upwards in the selection list Edit numeric value within a function
S or W Navigate downwards in the selection list Edit numeric value within a function
F or M Navigate to the right within a function group, confirmation.
or S and F
Contrast settings of the LCD
Hardware lock / unlock After a hardware lock, an operation of the instrument via display or communication is not possible! The hardware can only be unlocked via the display. An unlock parameter must be entered to do so.
Custody locking switch Access to the electronics can be prevented by means of a custody locking switch that locks the device settings; see fig. 4 on Page 34. The custody locking switch can be sealed for custody transfer applications.
Software reliability The software used in the 7500 RTG fulfills the requirements of OIML R85. This particularly includes:
The 7500 RTG continuously monitor the compliance with accuracy requirements for custody transfer measurements according to OIML R85. If the accuracy cannot be maintained, a specific alarm is generated on the local display and via the digital communication
Key(s) Meaning

The 7500 RTG can be protected in two ways against unauthorized changing of instru- ment data, numerical values or factory settings:
"unlock parameter" (0A4): A value <> 100 (e.g. 99) must be entered in "unlock parameter" (0A4) in the "diagnostics" (0A) function group. The lock is shown on the display by the symbol and can be released again either via the display or by communication.
Hardware lock: The instrument is locked by pressing the O and S and F keys at the same time. The lock is shown on the display by the symbol and can only be unlocked again via the display by pressing the O and S and F keys at the same time again. It is not possible to unlock the hardware by communication. All parameters can de displayed even if the instrument is locked.

⇓ O and S and F press simultaneous


The LOCK_SYMBOL appears on the LCD.
7532 Operation

Operation Radar Tank Gauge
If an attempt is made to change parameters when the instrument is locked, the user is automatically requested to unlock the instrument:
100 = for HART devices
the 7500 RTG is released for operation.
Hardware lock: After pressing the O and S and F keys at the same time, the user is asked to enter the unlock parameter
100 = for HART devices.
⇓

Please enter unlock code and confirm with F.

A reset is only necessary:

User input ("reset" (0A3)):
333 = reset customer parameters This reset is recommended whenever an instrument with an unknown 'history' is to be used in an application:
List of functions that are affected by a reset:
555 = History Reset After mounting and aligning the equipment, carry out a history reset before switching on the function "Auto-correction" (031) (see page 57).
The tank map can also be reset in the "cust. tank map" (055) function of the "extended calibr." (05) function group. This reset is recommended whenever an instrument with an unknown 'history' is to be used in an application or if a faulty mapping was started:
7532 Operation

Operation Radar Tank Gauge
Type of error Errors that occur during commissioning or measuring are displayed immediately on the local display. If two or more system or process errors occur, the error with the highest priority is the one shown on the display. The measuring system distinguishes between two types of error:

Error messages Error messages appear as four lines of plain text on the display. In addition, a unique error code is also output. A description of the error codes is given on page 79.

Apart from local operation, you can also parameterize the measuring instrument and view measured values by means of a HART protocol. There are two options available for operation:
All device functions can be adjusted via menu operation with the handheld unit DXR 375.

The ToF Tool is a graphical operating software for instruments from Varec that operate based on the time-of-flight principle. It is used to support commissioning, securing of data, signal analysis and documentation of the instruments. It is compatible with the following operating systems: Win95, Win98, WinNT4.0, Win2000 and Windows XP. The ToF Tool supports the following functions:
7532 Operation

Operation Radar Tank Gauge

Signal analysis via envelope curve

Connection options:
Menu-guided commissioning

Commuwin II is an operating software with graphical support for intelligent transmitters with the communication protocols Rackbus, Rackbus RS 485, INTENSOR, HART or PROFIBUS-PA. It is compatible with the operating systems Win 3.1/3.11, Win95, Win98 and WinNT4.0. All functions of Commuwin II are supported. The configuration is made via operating matrix or graphic surface. A envelope curve can be displayed in ToF Tool.
Connection The table provides an overview of the Commuwin connections.
Interface Hardware Server Device list
HART Commubox FXA 191 to HART Computer with RS-232C interface
Interface FXN 672 Gateway for MODBUS, PROFIBUS, FIP, INTERBUS, etc. Computer with RS-232C interface or PROFIBUS card
ZA 673 for PROFIBUS ZA 672 for other
List of all rack bus modules: the required FXN 672 must be selected
7532 Operation · HART Connected instrument

Operation Radar Tank Gauge

6.1 Function check
Make sure that all final checks have been completed before you start up your measuring point:
When the instrument is switched on for the first time, the following messages appear on the display:

⇓
After 5 s, the following message appears

After 5 s or after you have pressed F the follow- ing message appears

Select the language (this message appears the first time the instrument is switched on) ⇓

Select the basic unit (this message appears the first time the instrument is switched on)

The current measured value is displayed
After F is pressed, you reach the group selection.

This selection enables you to perform the basic setup
7532 Commissioning

Commissioning Radar Tank Gauge


settings in 005 settings in 006 D = distance (distance flange / product) L display in 0A5 in 0A6
display SD = safety settings BD = blocking dist. settings in 015 settings in 059
7532 Commissioning

Commissioning Radar Tank Gauge
To successfully commission a precise measurement to the nearest mm, it is important you carry out a history reset on first installation after mechanical installation and after the basic setup of the device (see Page 56). Only after a history reset the mounting cali- bration is carried out. Enter the measurement offset as the first point in the dip table for the mounting calibration. When a value is dipped at a later date, make a second entry into the dip table, again using the semi-automatic mode. This way, you can easily carry out a linearization of the measurement. When configuring the function in "basic setup" (00) please take into account the following notes:

Function "measured value" (000)
This function displays the current measured value in the selected unit (see "customer unit" (042) function). The number of digits after decimal point can be selected in the "no.of decimals" (095) function. The length of the bargraph corresponds to the percentile value of the present measured value with regard to the span.
⇒
Function "tank shape" (002)

This function is used to select the tank shape.
Selection:

7532 Commissioning

Commissioning Radar Tank Gauge
Function "medium property" (003)

This function is used to select the dielectric constant.
Selection:
A 1,4 … 1,9 non-conducting liquids, e.g. liquefied gas 3)
B 1,9 … 4 non-conducting liquids, e.g. benzene, oil, toluene, …
C 4 … 10 e.g. concentrated acids, organic solvents, esters, aniline, alcohol, acetone, …
D > 10 conducting liquids, e.g. aqueous solutions, dilute acids and alkalis
3)Treat Ammonia NH3 as a medium of group A, i.e. always use a stilling well.

Function "process cond." (004)

This function is used to select the process conditions.
Selection:
For all applications that do not fit into any of the following groups.
Storage tanks with immersion tube or bottom filling
The filter and output damping are set to average values.
The averaging filters and output damping are set to high values. -> steady meas. value -> precise measurement -> slower reaction time
7532 Commissioning · standard calm surface

Commissioning Radar Tank Gauge
Function "empty calibr." (005)

This function is used to enter the distance from the flange (reference point of the measurement) to the minimum level (=zero).

Function "full calibr." (006)

This function is used to enter the distance from the minimum level to the maximum level (=span).

In principle, it is possible to measure up to the tip of the antenna. However, due to considerations regarding corrosion and build-up, the end of the measuring range should not be chosen any closer than 50 mm (2”) to the tip of the antenna.

Function "pipe diameter" (007)

This function is used to enter the pipe diameter of the stilling well or bypass pipe.

Microwaves propagate slower in pipes than in free space. This effect depends on the inside diameter of the pipe and is automatically taken into account by the 7500 RTG. It is only necessary to enter the pipe diameter for applications in a bypass or stilling well. If mounting the 7532 RTG on stilling wells with a widening of the pipe, the inner diam- eter of the lower part of the pipe (d2 in the Fig.) must be entered. This is the part of the stilling well, where the measurement is actually performed.
Function "dist./ meas. value (008)"

The distance measured from the reference point to the product surface and the level calculated with the aid of the empty adjustment are displayed. Check whether the values correspond to the actual level or the actual distance. The following cases can occur:
7532 Commissioning

Commissioning Radar Tank Gauge
Function "check distance" (051)

This function triggers the mapping of interference echoes. To do so, the measured distance must be compared with the actual distance to the product surface. The following options are available for selection: Selection:

dist. too small
dist. too big
dist. unknown If the actual distance is not known, no mapping can be carried out.
manual A mapping is also possible by manual entry of the range to be suppressed. This entry is made in the "range of mapping (052)" function.
distance = ok

Function "range of mapping" (052)

This function displays the suggested range of mapping. The reference point is always the reference point of the measurement (see page 46). This value can be edited by the operator. For manual mapping, the default value is: 0 m.
Function "start mapping" (053)

This function is used to start the interference echo mapping up to the distance given in "range of mapping" (052).
Selection:
7532 Commissioning

Commissioning Radar Tank Gauge
Display "dist./meas.value (008)"
The distance measured from the reference point to the product surface and the level calculated with the aid of the empty alignment are displayed again. Check whether the values correspond to the actual level or the actual distance. The following cases can occur:
Function "history reset" (009)

By this function a history reset of the device is performed, i.e. the correspondence table between level an index values is deleted. A new correspondence table will be filled and stored after the history reset, see page 57.

⇓
After 3 s, the following message appears

⇒

Function "tank gauging" (030)

Using this function, you can either enter a dip table or carry out an auto-correction.

When measuring levels with radar systems, so-called "multipath reflections" can affect the level signal giving rise to serious measuring errors. "Multipath reflections" also include radar beams which are received by the radar system, which have not been reflected directly by the medium surface. They may reach the antenna via the basin wall and the medium surface. This phenomenon is particularly noticeable with devices mounted near to walls, as soon as the conical radar beam strikes the basin wall. The 7500 RTG can automatically discover and correct measuring errors due to this "multiple path" propagation. This is because it uses two independent sets of information when evaluating reflection signals:
The decisive factor for a constant output signal is to assign the phase values to the associated level values. This assignment is ensured using a correspondence table (index correction table). The 7500 RTG learns this for the specific application after installation (learning period). Therefore, after mounting the device, and after completing the basic calibration, a history reset must be performed (enter "yes" in the "history reset" (009) function in the "basic setup " (00) function group). Do not switch off the radar system during filling and emptying operations during the teach-in phase. Switching off when there are only negligible level changes produces no error. "
Caution! During the learning period, fast filling/emptying or turbulent surfaces can result in switching off and on the phase evaluation. Subsequently observed measurement errors will disappear as soon as tank levels come back to areas measured by 7500 RTG previously with activated phase evaluation. If the observed measurement errors are curetted by dip table entries, the 7500 RTG will take care of these corrections and automatically adjust the index correction table. Do NOT correct any settings in the basic calibration or the extended calibration.
7532 Commissioning · Function "auto-correction" (031)

Commissioning Radar Tank Gauge
correction table), the maximum permissible filling speed is 100 mm level change / min. After this, the fill speed has no limitation.
Function "pipe diam. corr." (032)

For level measurement in stilling wells, radar systems require highly precise pipe inner diameter data. An mm-exact level measurement cannot be guaranteed for deviations from the actual stilling well inner diameter of more than ± 0.1mm to the value entered in the function group "basic setup" (00). The errors which occur as a result are linear and can be corrected with a dip table containing at least two entries. The 7500 RTG also has an automatic pipe inner diameter correction. This adjusts the entered stilling well inner diameter (input in the function group "basic setup" (00)) to the actual values. However, this presupposes that the value entered in the function group "basic setup" (00) matches the actual pipe inner diameter accurately as possible. The user-defined value entered in the function group "basic setup" (00) can be corrected with this value. In order to do this, switch on the "pipe diam. corr." (032) function, after a level change of at least 5 m has occurred since start- up. The pipe diameter, which the instrument determines automatically, will then be transmitted to the "pipe diameter" (007) function.
Note! Only if the "pipe diameter" (007) function has changed its value, it is necessary to per- form a "history reset" (009) and to delete the dip table after activation of the "pipe diam. corr." (032) function. Otherwise the level change of 5 m has not yet been exceeded. The "pipe diam. corr." (032) function must be deactivated again and the procedure should be repeated at a later point of time.

Display "custody mode" (0A9)

This indicates the instrument calibration mode. The calibration mode (active) can be set using the hardware security lock on the electronics (see Page 34).
Selection:
active pos. The custody mode (instrument is lead-sealed and accurate to the nearest mm) is active and is held.
active neg. Custody mode (instrument is lead-sealed and accurate to the nearest mm) is activated and not held, e.g. because the signal-to-noise ratio is less than 10 dB (refer to function "echo quality" (056) in the function group "extended calibr." (05)).
7532 Commissioning

Commissioning Radar Tank Gauge
Dip table The dip table is used to correct the level readings of the 7500 RTG using independently taken hand dips. The dip table is used in particular to adapt the level gauge to the specific application conditions as mechanical offset and tank/stilling well design. Depending on national regulations, national inspectors will dip the tank at one to three levels during a calibration run and check the level readings. Only one value pair must be entered into the dip table to correct the measurement offset. If a second value pair is entered into the dip table, the 7500 RTG accepts the corrected measured values identically for both value pairs. All other measured values are determined by linear extrapolation. If you enter more than two value pairs, the system carries out a linear interpolation between adjacent value pairs. Outside these value pairs, extrapolation is also linear.
hand dip
collect hand dip and measurement data over full measurement range
collect further hand dip data; if deviation from measurements > 4 mm and level change > 2 m:
evaluate collected data; enter characteristic value
enter a new dip value into dip table using the
pairs into the dip table using the manual mode
Figure 8: Alternative procedures to fill the dip table.
To collect and enter data into the dip table, two alternative procedures may be carried out. In order not to mix up measurement values corrected by the offset or linearization of the dip table with uncorrected measurement values, it is recommended to use the semi-automatic mode of the dip table to enter new data pairs. In this case, the first dip value should be entered immediately after the basic calibration. Further linearization points should be entered only after a level change of at least 2 m (see Figure 8:, preferred choice) and a deviation between the "uncorrected measurement value" and the hand dip value of at least 4mm. If this procedure can not be followed, then NO value pair should be entered into the dip table after basic calibration. Measurement data and hand dip values should be collected over the full measurement range and be evaluated with regard to a good linear fit. Only then characteristic value pairs should be entered into the dip table using the "manual mode" (see Figure 8:, right side). If further linearization is needed, further hand dip values should be entered using only the "semi-automatic mode".

Function "dip table state" (037)

This function displays the dip table status.
Display:
table on Indicates whether the dip table is active.
table off Indicates whether the dip table is not active.
Function "dip table mode" (033)

The dip table can be switched on or off using this function.
Selection:
7532 Commissioning

Commissioning Radar Tank Gauge
manual The value pairs in the dip table can be read and written. You can enter the measured value and the dip value.
–uncorrected measured value:
This is the measured value supplied by the instrument, NOT corrected by the dip table. The choice of measured value, level or remaining fill height is dependent on the instrument setting. –Dip value:
This is the level or distance to flange respectively, given by the hand dip. This value should be used to correct the measured value. The "manual mode" of the dip table can be used to enter collected data after a series of data pairs taken at different tank levels.
semi-automatic The value pairs in the dip table can be read. You can enter the dip value only. When there are new value pairs, the current level or distance is accepted as the measured value.
table on The dip table is switched on.
table off The dip table is switched off.
clear table The complete dip table is deleted. The table is switched off. The number of free table entries is set to the maximum value (= 32).
View The value pairs in the dip table can only be read. You can still select this menu option, even if there is no dip table available. In this case, the number of free table entries is at maximum value (= 32).
Function "dip table" (034)

This function edits measured variable. The number behind the entry "remain" indicates the current number of remaining free value pairs. The maximum number of value pairs is 32; after each entry, the remaining number is decremented.
Function "dip table" (035)

This function edits the dip value.

Function "dip table handl." (036)

Use this function to enter the dip value (level or distance) which will correct the mea- surement values.
Selection:
General procedure: To enter a new point into the dip table, use "new point" to enter the value (pairs), "store point" to sort the new value (pairs), "return" to go to the dip table mode and "table on" to activate the dip table.
new point You can enter a new point. A suggested value is displayed for the measured variable, the dip value of the current level or the distance. With semi-automatic inputs, the current level or the current remaining fill height is displayed as proposed dip value. The new value pair can be altered without selecting the "edit point" parameter. If the table is full, you can still select this parameter. In this case, the number of free table entries stands at minimum value (= 0).
edit point The displayed value pair can be changed. Only the dip value can be changed with semi- automatic input mode.
store point The displayed value pair is sorted in the table.
7532 Commissioning

Commissioning Radar Tank Gauge
delete point The currently displayed point is deleted from the table. After deletion, the previous point is displayed. If the table only consisted of one point before deletion, then the current measured variable is displayed as a value pair.
return By selecting this point, you return to the function "dip table mode" (033).
next point This scrolls down in the table. If the table is empty, you can still select this option. However, the displayed value does not change.
previous point This scrolls up in the table. If the table is empty, you can still select this option. However, the displayed value does not change.

After the basic setup, an evaluation of the measurement with the aid of the envelope curve ("display" (09) function group) is recommended.
Function "plot settings" (09A)

Here you can select which information is shown on the display:

Function "recording curve" (09B) This function determines whether the envelope curve is read as:

7532 Commissioning

Commissioning Radar Tank Gauge
To carry out the basic setup with the ToF Tool operating program, proceed as follows:
The following display appears on the screen:
Basic Setup step 1/5:

Change to the next page

Basic Setup step 2/5:

Basic Setup step 3/5: If "stilling well" or "bypass" is selected in the "tank shape" function, the following display appears on the screen:

7532 Commissioning

Commissioning Radar Tank Gauge
Basic Setup step 4/5:

Step 5/5: After the first installation of the device, initialise the index correction table (compare page 57) by activating the history reset 555.
measured value · distance

After the basic setup, an evaluation of the measurement using the envelope curve is recommended.

To set the parameters of user-specific applications, see the 7500 RTG "Service manual and description of instrument functions".
7532 Commissioning

Commissioning Radar Tank Gauge
To carry out the basic setup with the ToF Tool operating program, proceed as follows:
The following display appears on the screen:
Mounting calibration step 1/2:

change to next page

Mounting calibration step 2/2:

7532 Commissioning

Commissioning Radar Tank Gauge

The 7500 RTG measuring instrument requires no special maintenance.
Exterior cleaning When cleaning the exterior of measuring devices, always use cleaning agents that do not attack the surface of the housing and the seals.
Replacement After a complete 7500 RTG or electronic module has been replaced, the parameters can be downloaded into the instrument again via the communication interface. Prerequisite to this is that the data were uploaded to the PC beforehand using the ToF Tool / Commuwin II. Measurement can continue without having to carry out a new setup.
After an antenna component has been replaced, a new calibration must be carried out.
7532 Maintenance

Maintenance Radar Tank Gauge

Various accessories, which can be ordered separately from Varec, are available for the 7500 RTG.
Weather protection cover A Weather protection cover made of stainless steel is available for outdoor mounting (order code: 543199-0001). The shipment includes the protective cover and tension clamp.

Commubox FXA 191 HART For intrinsically safe communication with ToF Tool or Commuwin II via the RS 232C- interface.
Service adapter FXA 193 For communication with ToF Tool via the display connector. (Bestell-Nr.: 50095566).
Commuwin II Operating software for intelligent instruments.
7532 Accessories

Accessories Radar Tank Gauge

9.1 Troubleshooting instructions
7532 Troubleshooting

Troubleshooting Radar Tank Gauge


Code Description Possible cause Remedy
A102 checksum error general reset & new calibr.required
device has been powered off before data could be stored; emc problem; E²PROM defect
reset; avoid emc problem; if alarm prevails after reset, exchange electronics
W103 initialising - please wait
E²PROM storage not yet finished
wait some seconds; if warning prevails, exchange electronics
A106 downloading please wait
processing data download wait until warning disappears
A110 checksum error general reset & new calibr.required
device has been powered off before data could be stored; emc problem; E²PROM defect
reset; avoid emc problem; if alarm prevails after reset, exchange electronics
A111 electronics defect RAM defective reset; if alarm prevails after reset, exchange electronics
A113 electronics defect ROM defective reset; if alarm prevails after reset, exchange electronics
A114 electronics defect E2PROM defective reset; if alarm prevails after reset, exchange electronics
A115 electronics defect general hardware problem reset; if alarm prevails after reset, exchange electronics
A116 download error repeat download
checksum of stored data not correct
restart download of data
A121 electronics defect no factory calibration existant; E²PROM defective
W153 initialising - please wait
initialisation of electronics wait some seconds; if warning prevails, power off device and power on again
A155 electronics defect hardware problem reset; if alarm prevails after reset, exchange electronics
A160 checksum error general reset & new calibr.required
device has been powered off before data could be stored; emc problem; E²PROM defect
reset; avoid emc problem; if alarm prevails after reset, exchange electronics
A164 electronics defect hardware problem reset; if alarm prevails after reset, exchange electronics
A171 electronics defect hardware problem reset; if alarm prevails after reset, exchange electronics
A231 sensor 1 defect check connection
HF module or electronics defective
exchange HF module or electronics
7532 Troubleshooting · contact service

Troubleshooting Radar Tank Gauge
Code Description Possible cause Remedy
A270 custody switch undef check position
switch for custody transfer may be defective
check position of custody switch exchange electronics
# inconsistency between phase and amplitude evaluation inconsistent microfactor inconsistent index mapping
check basic calibration check mounting calibration check echo quality reset history "555" check stilling pipe diameter switch off autocorrection
A272 electronics defect amplifier
inconsistency in amplification exchange electronics
W275 electronics defect factory setting
offset drift of A/D commuter exchange electronics
W511 no factory calibration ch1
factory calibration has been deleted
record new factory calibration
A512 recording of mapping please wait
mapping active wait some seconds until alarm disappears
W601 linearization ch1 curve not monotone
linearization not monotonously increasing
W611 less than 2 linearization points for channel 1
number of entered linearization points < 2
W621 simulation ch. 1 on simulation mode is active switch off simulation mode
E641 no usable echo channel 1 check calibr.
echo lost due to application conditions of built up on antenna
check installation; optimize orientation of antenna; clean antenna
E651 level in safety distance - risk of overspill
level in safety distance alarm will disappear as soon as level leaves safety distance;
A671 linearization ch1 not complete, not usable
linearization table is in edit mode
W681 current ch1 out of range
current out of range (3,8 mA ... 21,5 mA)
check calibration and linearization
correct linearization table · correct linearization table · activate linearization table

Error Output Possible cause Remedy
A warning or alarm has occurred.
Depending on the configuration See table of error messages (see page 79)

yes → 1. Check empty calibr. (005) and full calibr. (006).
yes → 1. Is bypass or stilling well selected in tank shape (002)?
yes → 1. offset (057) correctly set?
yes → 1. Carry out tank mapping → basic setup


Interference echo from installations, nozzle or extension on the antenna
7532 Troubleshooting

Troubleshooting Radar Tank Gauge


If the surface is not calm (e.g. filling, emptying, agi- tator run- ning), the measured value jumps sporad- ically to a higher level
Signal is weakened by the rough surface — the interference echoes are sometimes stronger

Multiple echoes yes → 1. Check the tank shape (002), e.g. "dome ceiling" or "horizontal cyl"



yes → 1. Check application parameters (002), (003) and (004)

E 641 (loss of echo) after turn on the power supply
If the instrument is configured to Hold by loss of echo the output is set to any value/current.
Repeat once more empty calibr. (005). Caution! Before conformation change with O or S to the edit mode.

7532 Troubleshooting

Troubleshooting Radar Tank Gauge

The customer may exchange modular assemblies in accordance with Varec instructions!

Spare parts 7532 RTG Planar antenna Process connection without antenna
detailed order structure and variants see spare parts price list
O-ring 49,21 x 3,53 EPDM (Seal set 52005628)
O-ring 49,21 x 3,53 EPDM (Seal set 52005628)
O-ring 85,32 x 3,53 FKM (green) 85,32 x 3,53 HNBR12-70 (black)
O-ring 85,32 x 3,53 FKM (green) 85,32 x 3,53 HNBR12-70 (black)
5200 6193 DN 150 ( 6") HNBR 5200 6332 DN 150 ( 6") FKM 5200 6194 DN 200 ( 8") HNBR 5200 6333 DN 200 ( 8") FKM 5200 6195 DN 250 (10") HNBR 5200 6334 DN 250 (10") FKM
Modification nameplate When ordering parts that are listed in the product structure (see page 8) , a check must be made as to whether the instrument description on the nameplate is still valid, e.g. for:
7532 Troubleshooting · L00-FMR532xx-00-00-06-en-004 · Antenna size Seal

Troubleshooting Radar Tank Gauge
Housing T12 The complete order code must be specified when ordering the replacement housing so that the correct nameplate can be delivered, e.g.
You must label the nameplate yourself.
If you need to send a 7500 RTG back to Varec for repair, please send a completed copy of the form printed on the last page.
Please carry out the measures described in section 1.4, “Return” on page 5 before sending a 7500 RTG back for repair.
In case of disposal please separate the different components according to their material consistence.
Software version / Date
Software changes Documentation changes
V 1.00.00 / 12.2000 Original software. Operated via:
– ToF Tool – Commuwin II (from version
2.05.03) – HART communicator DXR 375
with Rev. 1, DD 1.
V 01.02.00 / 03.2002 simplified commissioning history reset in basic calibration. Operated via:
– ToF Tool (V3.0) – Commuwin II (from version
2.05.03) – HART communicator DXR 375
with Rev. 1, DD 1.

10.1 Technical data at a glance
Application The 7500 RTG is used for highly accurate level measurement in storage tanks and can be applied in custody transfer applications. It meets the relevant requirements according to OIML R85 and API 3.1B. • The 7532 RTG with planar antenna is specifically and only suited for stilling well applications with ranges up to 38 m (124 ft).
Measuring principle The 7500 RTG is a "downward-looking" measuring system, operating based on the time-of-flight method. It measures the distance from the reference point (process connection) to the product surface. Radar impulses are emitted by an antenna, reflected off the product surface and received again by the radar system.
Measured variable The measured variable is the distance between a reference point (mounting flange) and a reflective surface (e.g. medium surface). The measured value and all parameters are displayed using either metrical SI-units or US/UK-units (inch, ft, …). The level is calculated based on the tank height entered. The level can be converted into other units (volume, mass) by means of a linearization. In order to compensate for non-linear effects like movement of the tank roof, an additional correction table (diptable) can be entered.
Measuring range See page 15
Output signal 4…20 mA with HART protocol: this version can be operated via the PC operating software ToF Tool and Commuwin II. The instrument supports both point-to-point and multidrop operation.
Signal on alarm Error information can be accessed via the following interfaces: • Local display: – Error symbol (see page 35) – Plain text display – LED’s: red LED continuously on = alarm, red LED flashes = warn-
Galvanic isolation 500 V towards ground. 500 V between power supply and signal
7532 Technical data

Technical data Radar Tank Gauge
Electrical connection Housing T 12 with separate terminal compartment.
Load HART Minimum load for HART communication: 250 Ω
Supply voltage See page 28
Power consumption Max. 330 mW at 16 V, max. 500 mW at 24 V, max. 600 mW at 30 V.
Current consumption Max. 21 mA (50 mA inrush current).
Power supply For stand alone operation recommended via e.g. Varec RN 221N/Z.
The accuracy of each 7500 RTG is established through a calibration certificate that records the absolute and relative error at 10 equidistant points during the final test. For stilling well measurements with 7532 RTG, a NMI / PTB calibrated tape with an absolute accuracy of 0.25 mm is used. Each 7500 RTG is delivered with the PTB and NMi type approval. Additional initial factory verifications for custody applications are available on demand for all 7500 RTG.
Proof of accuracy for custody transfer ver- sions
Maximum fill speed By the first pass trough of measuring range: 100 mm/min., after it unlimited.
Non-repeatability 0,3 mm (1/64")
Hysteresis 0.3 mm
Settling time Typical 15 sec
Long-term drift The long-term drift is within the specified accuracy.
Influence of ambient temperature
Within the specified accuracy according to OIML R85.
Reference operating conditions · Maximum measured error

All device types can be delivered as "Inventory Control Versions" with a reduced accuracy of ± 3mm (under reference conditions). To these versions, the calibration certificate or custody transfer type approval is NOT attached. The "Inventory Control Versions" can be selected by choosing the option »R« in feature »70« in the order code section »Custody transfer approvals« on page 8.
Engineering tips See page 14
Storage temperature -40 °C ... +80 °C (-40 °F … +176°F)
Climate class DIN EN 60068-2-38 (test Z/AD)
Vibration resistance DIN EN 60068-2-64 / IEC 68-2-64: 20…2000 Hz, 5 (m/s²)²/Hz
Cleaning of the antenna See »Technical Information« TEC028
See »Technical Information« TEC028
See »Technical Information« TEC028
Process pressure limits See »Technical Information« TEC028

Technical data Radar Tank Gauge
Design, dimensions See page 13
Weight Approx 6 kg + weight of flange
Material See page 8
Process connection see page 8 All process connections dispose of a gas-tight glass feed-through to prevent any gas leakage to the inside of the housing.
Operation concept See page 31
Display See page 31
CE approval The measuring system meets the legal requirements of the EC- guidelines. Varec confirms the instrument passing the required tests by attaching the CE-mark.
RF approvals R&TTE 1999/5/EG, FCC CRF 47, part 15
Overspill protection PTB , NMi and many other national approvals
External standards and guidelines
EN 60529 Protection c3lass of housing (IP-code) EN 61010 Safety regulations for electrical devices for measurement, control, regulation and laboratory use. EN 61326 Emissions (equipment class B), compatibility (appendix A – industrial area) NAMUR Standards committee for measurement and control in the chemical industry API (American Petroleum Institute) Particularly "Manual of Petroleum Measurement Standards". OIML R85 (Organisation Internationale de Metrologie Legale)
Ex approval XA 081F-A Installation 7500 RTG (T12 / EEx ia IIC T6…T1) PTB 00 ATEX 2067 X, Equipment marking: (II 1/2 G)
The Varec service organisation can provide detailed ordering information an information on the order codes on request.
Supplementary Documentation
see page 75 · Supplementary Documentation

11.1 Operating menu HART (Display modul), ToF Tool
7532 Appendix

Appendix Radar Tank Gauge

dome ceiling horizontal cyl. bypass stilling well flat ceiling sphere
standard calm surface turb. surface add. agitator fast change test: no filter
in case of echo loss max. 4000 sec. default: 30 s
next point previous point new point edit point store point delete point
distance = ok dist. too small manual dist. unknown dist. too big
custody transfer Note! The default values of the parameters are typed in boldface.


distance = ok dist. too small dist. too big manual
7532 Appendix

Appendix Radar Tank Gauge

Function group V-CWII H0 H1 H2 H3 H4 H5 H6 H7 H8 H9
05 extended calibr. V4 check distance range of mapping start mapping pres. Map dist cust. Tank map echo quality offset output damping blocking dist.
0A diagnostics V9 present error previous error clear last error reset unlock parameter measured dist. measured level application par. custody mode
00 basic setup V0 measured value tank shape medium property process cond. empty calibr. full calibr. pipe diameter
01 safety settings V1 output on alarm output on alarm outp. echo loss ramp %span/min delay time safety distance in safety dist. ackn. alarm overspill protection
04 linearisation V3 level/ullage linearisation customer unit table no. input level input volume max. scale diameter vessel
06 output V5 commun. Address no. Of preambels low output limit fixed current fixed current simulation simulation value output current
09 display V6 language back to home format display no. of decimals sep. character
0C system parameter VA tag no. protocol+sw-no. software no. serial no. distance unit download mode
03 mounting calibr. V2 tank gauging auto correction pipe diam. corr. dip table mode dip table dip table dip table handl. dip table state
7532 Appendix · 0D service V7

Appendix Radar Tank Gauge

The 7500 RTG is a "downward-looking" measuring system, operating based on the time-of-flight method. It measures the distance from the reference point (process connection) to the product surface. Radar impulses are emitted by an antenna, reflected off the product surface and received again by the radar system.

Input The reflected radar impulses are received by the antenna and transmitted into the electronics. A microprocessor evaluates the signal and identifies the level echo caused by the reflection of the radar impulse at the product surface. The unambiguous signal identification is accomplished by the PulseMaster® software based on many years of experience with time-of-flight technology. The mm-accuracy of the 7500 RTG is achieved with the patented algorithms of the PhaseMaster® software. The distance D to the product surface is proportional to the travel time t of the impulse:
D = c · t/2, with c being the speed of light.
Based on the known empty distance E, the level L is calculated:
Reference point for "E" is the lower surface of the process connection. The 7500 RTG is equipped with functions to suppress interference echoes. The user can activate these functions. They ensure that interference echoes (e.g. from edges and weld seams) are not interpreted as level echo.
7532 Appendix

Appendix Radar Tank Gauge
Output The 7500 RTG is commissioned by entering an empty distance E (=zero), a full distance F (=span), and an application parameter. The application parameter automatically adapts the instrument to the measuring conditions. The data points “E” and “F” correspond with 4mA and 20mA for instruments with current output. They correspond with 0 % and 100% for digital outputs and the display module. A linearization with max. 32 points, based on a table entered either manually or semi- automatically, can be activated locally or remotely. This function provides a measurement in engineering units and a linear output signal for spheres, horizontal cylindrical tanks, and vessels with conical outlet.
Stand-alone The 7500 RTG can be used for measurement in a stilling well as well as in free space. The different instrument versions are applied as follows:
The complete measuring system consists of:


On-site operation:
Remote operation:

Integration into the Asset Management System The HART interface allows the integration into the AMS® (Asset Management System) from Fisher-Rosemount.
7532 Appendix

Appendix Radar Tank Gauge
7500 RTG is a weight and measure approved level transmitter. Either the innage or the ullage can be selected as the custody transfer variable.
The selected variable is the basis for the subsequent calculation of the current amount of product in a tank, along with other measured variables such as (average) temperature and pressure.
This opens up numerous application options in custody transfer:
The type approvals for custody transfer issued by the PTB and NMi, a copy of which is enclosed with every device, prove the fundamental suitability of the various types for custody transfer.
In addition to this, the accuracy of every single device is documented using a calibration certificate, which is issued in the factory after the device has been tested on a reference test rig.
On request, a separate initial verification of the devices can be carried out with a National Standards Authorities inspector present, who issues a preliminary test certifi- cate for every device. In the initial verification, the device is tested to ensure that it com- plies with the limit of error in legal metrology, which lies at +-2mm for radar measuring devices in Germany.
Essentially, this proves that the devices are weight and measure approved. The devices must not, however, be used in custody transfer mode straight away.
The measuring device is not approved until after the approval after installation by the Standards Authorities. For this, the device's level measurement is compared with the tank gauging by a National Standards Authorities inspector using manual dips (also "Ini- tial verification"). As a rule, a quiescent tank gauging is dipped by hand three times in a row and then compared with the value displayed by the level radar. Depending on national regulations, the transfer error limit, calculated as the arithmetic mean of the absolute deviations of all three measurements, must not exceed double the limit of error in legal metrology (compare, for example, the German "Eichordnung" or the American "API 3.1B", in which the necessary procedures are also defined).
Depending on national regulations, this test is repeated with various tank gaugings. Using linearization tables to compensate any non-linearities that occur in measurement is permitted. For this, the 7500 RTG level radar offers a special dip table, compare “Mounting calibration with VU 331” on page 57.
After the measurement has been approved by an inspector, he seals the level radar at the stamp position and thereby also secures the programming status of the device.
Those operating an approved level transmitter are obligated to obtain reapproval in accordance with the applicable national regulations from the Standards Authorities.

The 7500 RTG level radar is set to custody transfer mode after commissioning using a custody locking switch (see page 36). The position of the custody locking switch is secured and sealed using the sealing pin.
During custody transfer measurement, all custody transfer-relevant functions for operation are automatically locked, so that the device software can not be used, either via local operation or via digital communication settings. This locked status is displayed by the key symbol ( ).
7500 RTG radar devices continuously monitor the compliance with accuracy requirements for custody transfer measurements according to OIML R85. If, for example, the accuracy cannot be maintained due to quick surface movements, this is reported via a separate alarm in the local display (displays "#" symbol) and via digital communication.
For definitions and procedures please refer to the following documents:
7532 Appendix

Appendix Radar Tank Gauge
The Varec Tank Side Monitor 4590 provides integrated communications for sites with multiple tanks, each with one or more sensors on the tank, such as radar, spot or average temperature, capacitive probe for water detection and/or pressure sensors. Multiple protocols out of the Tank Side Monitor guarantee connectivity to nearly any of the existing industry standard tank gauging protocols. Optional connectivity of analog 4...20 mA sensors, digital I/O and analog output simplify full tank sensor integration. Use of the proven concept of the intrinsically safe HART bus for all on-tank sensors yields extremely low wiring costs, while at the same time providing maximum safety, reliability and data availability.

This product may be protected by at least one of the following patents.
Further patents are pending.





Varec, Inc. 5834 Peachtree Corners East, Norcross (Atlanta), GA 30092 USA Tel: +1 (770) 447-9202 Fax: +1 (770) 662-8939
Radar Tank Gauge · Your offical representative
What the fault codes shown on the display mean and how to clear them.
| Code | Meaning | Fix | |
|---|---|---|---|
| W103 | initialising - please wait | wait some seconds; if warning prevails, exchange electronics | p.79 |
| W153 | initialising - please wait | wait some seconds; if warning prevails, power off device and power on again | p.79 |
| W275 | electronics defect factory setting | exchange electronics | p.80 |
| W511 | no factory calibration ch1 | record new factory calibration | p.80 |
| W601 | linearization ch1 curve not monotone | correct linearization table | p.80 |
| W611 | less than 2 linearization points for channel 1 | correct linearization table | p.80 |
| W621 | simulation ch. 1 on | switch off simulation mode | p.80 |
| E641 | no usable echo channel 1 check calibr. | check installation; optimize orientation of antenna; clean antenna | p.80 |
| E651 | level in safety distance - risk of overspill | alarm will disappear as soon as level leaves safety distance; | p.80 |
| W681 | current ch1 out of range | check calibration and linearization | p.80 |
Owners and our team usually reply within a day — and your question helps the next person too.
Be the first to ask — other owners or our team will answer.