Page 1 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 Leica HER2 FISH System - 30 T est Instructions For Use For use on Leica Biosystems’ BOND-MAX and BOND-III System.
English Page 2 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 Contents Intended Use .................................................................................................
Page 3 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English Intended Use For in vitro diagnostic use The Leica HER2 FISH System .
English Page 4 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 The characteristic morphology of the tissue and the spatial distribution of oncogene copies in individual uncultured primary breast carcinomas are retained.
Page 5 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English The Leica HER2 FISH System - 30 T est is for use only on the Leica BOND-MAX and BOND-III System.
English Page 6 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 T issue sections mounted on charged slides (Leica BOND Plus Slides S21.21 13) can be held for up to 12 months at 2–8 °C before staining.
Page 7 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English bandpass uorescence microscope lter sets optimized for use with the Leica HER2 FISH System - 30 T est are available for most microscope models.
English Page 8 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 5. T urn on the PC attached to the Leica BOND-MAX and BOND-III System.
Page 9 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English Signal Assessment and Enumeration T o assess the signal quality and enumerate the HER2 and CEP17 signals follow the process below: 1.
English Page 10 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 Recommended Method for LSI HER2 to CEP17 Ratio Determination T o determine the LSI HER2 to CEP17 ratio, use the following method: 1.
Page 1 1 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English Figure 1: Interpretation Guide Leica HER2 FISH S ystem - 30 T est Interpretation Guide Count as 2 orange signals and 1 green signal Do not count.
English Page 12 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 Leica HER2 FISH S ystem - 30 T est Score Sheet 20 Nuclei Signal Cou.
Page 13 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English Quality Control Use of Control Slides It is recommended that a Leica HER2 FISH Control Slide is included in each test run to monitor assay performance and to assess the accuracy of signal enumeration.
English Page 14 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 Unexpected FISH staining, or variations in the staining, may be a result of alterations in the expression levels of the encoding genes.
Page 15 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English 2x2 Concordance Results Leica BOND-MAX System Data was grouped as negative (<2.
English Page 16 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 2x2 Concordance Results Leica BOND-III System Data was grouped as negative (<2.
Page 17 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English Precision T esting – Leica BOND-MAX System A. Within Run Precision Study The within run precision study was performed in a randomized and blinded fashion.
English Page 18 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 F . Lot-to-Lot Precision Study The lot-to-lot precision study was performed in a randomized and blinded fashion.
Page 19 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English K. Between Observer Precision Study The between observer precision study was performed in a randomized and blinded fashion.
English Page 20 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 denaturation time is recommended.
Page 21 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English T roubleshooting Problem Probable C ause Remedial Action No or weak.
English Page 22 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 Problem Probable C ause Remedial Action Nonspecic background uorescent signal/staining Inadequate post hybridization washing Increase post hybridization wash incubation time.
Page 23 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English References 1. Bargmann CI, Hung MC, Weinberg RA. The neu oncogene encodes an epidermal growth factor related protein. Nature 1986;319:226–30.
Page 24 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 License Agreement This product contains PathV ysion FISH probes supplied by Abbott Molecular Inc. PathV ysion, LSI and CEP are a trademark of Abbott Molecular Inc.
An important point after buying a device Leica TA9217 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Leica TA9217 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Leica TA9217 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Leica TA9217 you will learn all the available features of the product, as well as information on its operation. The information that you get Leica TA9217 will certainly help you make a decision on the purchase.
If you already are a holder of Leica TA9217, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Leica TA9217.
However, one of the most important roles played by the user manual is to help in solving problems with Leica TA9217. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Leica TA9217 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center