EN
Invacare logo
Official user manual

Invacare XP6500

64 pages·2.1 MB·Mobility Aid·PDF·ENDEFRITNLJA
Invacare XP6500
Table of Contents 80 sections SPECIAL NOTES · NOTICE · TYPICAL PRODUCT PARAMETERS · LABEL LOCATIONS · … ShowHide
SPECIAL NOTESp.6NOTICEp.6TYPICAL PRODUCT PARAMETERSp.8LABEL LOCATIONSp.9 p.9DO NOT OPERATE WITHOUTp.9THE ANTI-TIP TUBESp.9INSTALLED. REV. 5/98 P/N 60106X144p.9SECTION 1—GENERAL GUIDELINESp.10Controller Settings/Repair or Servicep.10Operating Informationp.10Tire Pressurep.12Electrical Accessoriesp.12Batteriesp.12Grounding Instructionsp.14Rain Testp.14Weight Trainingp.15Weight Limitationp.15SECTION 2—EMI INFORMATIONp.16SECTION 3—SAFETY/HANDLING OF WHEELCHAIRSp.18Stability and Balancep.18Coping With Everyday Obstaclesp.18A Note to Wheelchair Assistantsp.19Tippingp.19Lifting/Stairwaysp.20Transferring to and From Other Seatsp.21Percentage of Weight Distributionp.22Reaching, Leaning and Bending - Forwardp.22Reaching and Bending - Backwardp.23SECTION 4—SAFETY INSPECTION/TROUBLESHOOTINGp.24Safety Inspection Checklistsp.24Troubleshooting - Mechanicalp.26X X X Xp.26X X X X Xp.26Troubleshooting Guidep.26SYMPTOM PROBABLE CAUSE SOLUTIONSp.26Checking Battery Charge Levelp.27DON’T DOp.27SECTION 5—WHEELCHAIR OPERATIONp.28Joystick Switches and Indicatorsp.28Operating the Wheelchairp.31MOVEMENT ACTIONp.32SECTION 6—FRONT RIGGINGSp.33Installing/Removing/Using the Footrest/Legrestp.33Adjusting the Footrest Heightp.34Replacing the Heel Loopp.35Raising/Lowering the Legrest Assemblyp.35SECTION 7—ARMSp.37Adjusting Armrest Heightp.37Swing-Back Armsp.38Replacing Armrest Padp.38SECTION 8—SEAT AND BACKp.39Replacing the Seat Upholsteryp.39SEAT DEPTH NUMBER OF SCREWSp.39Replacing the Back Upholsteryp.40Adjusting the Back Heightp.41*HOLE# 1 2 3p.41HEIGHT 17 18 19p.41Replacing Seat Positioning Strapp.42SECTION 9—JOYSTICKp.43Preparing the Joystick for Usep.43Repositioning the Joystickp.44SECTION 10—CASTERSp.45Adjusting Forksp.45SECTION 11—BATTERIESp.46When to Charge Batteriesp.46Charging Batteriesp.47TOOLS QUANTITY COMMENTSp.47Replacing the Batteriesp.48CHAIR SEATp.48QTY VOLTS BATTERYp.48REMARKSp.48WIDTHp.48SIZEp.48Removing/Installing the Battery Boxesp.49TOOL QTY COMMENTSp.51Connecting Battery Cablesp.52DO NOT USEp.52NEGATIVE (-)p.52SECTION 12—CLUTCH/MOTOR LOCKp.56
Page 1 of the Invacare XP6500 user manual
Page 1 · click to enlarge
Page 1
Invacare XP6500 diagram (page 1)
Fig. 1

DEALER: This manual MUST be given to the user of the wheelchair.

USER: BEFORE using this wheelchair, read this manual and save for future reference.

Page 2 of the Invacare XP6500 user manual
Page 2 · click to enlarge
Page 2

WARNING DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTANDING THIS MANUAL. IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS, CAUTIONS, AND INSTRUCTIONS, CONTACT A HEALTHCARE PROFESSIONAL, DEALER OR TECHNICAL PERSONNEL IF APPLICABLE BEFORE ATTEMPTING TO USE THIS EQUIPMENT - OTHERWISE INJURY OR DAMAGE MAY RESULT.

THE INITIAL SET UP OF THIS WHEELCHAIR MUST BE PERFORMED BY A QUALIFIED TECHNICIAN.

PROCEDURES OTHER THAN THOSE DESCRIBED IN THIS MANUAL MUST BE PERFORMED BY A QUALIFIED TECHNICIAN.

DEALERS AND QUALIFIED TECHNICIANS: DO NOT SERVICE OR OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTANDING (1) THE OWNER’S OPERATOR AND MAINTENANCE MANUAL, (2) THE SERVICE MANUAL (IF APPLICABLE) AND (3) THE SEATING SYSTEM’S MANUAL (IF APPLICABLE). IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS, CAUTIONS AND INSTRUCTIONS, CONTACT PROBASICS TECHNICAL SUPPORT BEFORE ATTEMPTING TO SERVICE OR OPERATE THIS EQUIPMENT - OTHERWISE, INJURY OR DAMAGE MAY RESULT.

Page 3 of the Invacare XP6500 user manual
Page 3 · click to enlarge
Page 3
Page 3 of the Invacare XP6500 user manual
Page 4 of the Invacare XP6500 user manual
Page 4 · click to enlarge
Page 4
Page 4 of the Invacare XP6500 user manual
Page 5 of the Invacare XP6500 user manual
Page 5 · click to enlarge
Page 5
Page 5 of the Invacare XP6500 user manual
Page 6 of the Invacare XP6500 user manual — NOTICE
Page 6 · click to enlarge
Page 6

SPECIAL NOTES

SPECIAL NOTES

Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the table below for definitions of the signal words.

SIGNAL WORDMEANING DANGERDanger indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury. WARNINGWarning indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. CAUTIONCaution indicates a potentially hazardous situation which, if not avoided, may result in property damage.

NOTICE

THE INFORMATION CONTAINED IN THIS DOCUMENT IS SUBJECT TO CHANGE WITHOUT NOTICE. WHEELCHAIR USER As a manufacturer of wheelchairs, ProBasics endeavors to supply a wide variety of wheelchairs to meet many needs of the end user. However, final selection of the type of wheelchair to be used by an individual rests solely with the user and his/her healthcare professional capable of making such a selection. WHEELCHAIR TIE-DOWN RESTRAINTS AND SEAT RESTRAINTS Wheelchair users should NOT be transported in vehicles of any kind while in wheelchairs. As of this date, the Department of Transportation has not approved any tie-down systems for transportation of a user while in a wheelchair, in a moving vehicle of any type. It is ProBasics’ position that users of wheelchairs should be transferred into appropriate seating in vehicles for transportation and use be made of the restraints made available by the auto industry. Invacare cannot and does not recommend any wheelchair transportation systems. With regard to restraints - seat/chest positioning straps - it is the obligation of the DME dealer, therapists and other healthcare professionals to determine if a seating positioning strap is required to ensure the safe operation of this equipment by the user. Serious injury can occur in the event of a fall from a wheelchair.

Page 7 of the Invacare XP6500 user manual — SPECIAL NOTES
Page 7 · click to enlarge
Page 7

SPECIAL NOTES

ProBasics products are specifically designed and manufactured for use in conjunction with ProBasics accessories. Accessories designed by other manufacturers have not been tested by ProBasics and are not recommended for use with ProBasics products. Wheelchairs should be examined during maintenance for signs of corrosion (water exposure, incontinence, etc.). Electrical components damaged by corrosion should be replaced IMMEDIATELY. Wheelchairs tat are used by incontinent users and/or are frequently exposed to water may require replacement of electrical components more frequently. The seat positioning strap is a positioning belt only. It is not designed for use as a safety device withstanding high stress loads such as auto or aircraft safety belts. If signs of wear appear, belt MUST be replaced immediately.

Page 8 of the Invacare XP6500 user manual — TYPICAL PRODUCT PARAMETERS
Page 8 · click to enlarge
Page 8

TYPICAL PRODUCT PARAMETERS

TYPICAL PRODUCT PARAMETERS

XP6500XP6520 SEAT WIDTH18 inches20 inches SEAT DEPTH16 or 18 inches (by extension)18 inches BACK HEIGHT17, 18, or 19 inches SEAT-TO-FLOOR19½ inches OVERALL WIDTH (EXCLUDING JOYSTICK)Unfolded - 24½ inches Folded - 13 inchesUnfolded - 26½ inches Folded - 13 inches OVERALL HEIGHT38 inches (with 17 inch back) OVERALL LENGTH28 inches (base only) to 47 inches (with riggings and anti-tippers) WEIGHT W/O BATTERIES AND FRONT RIGGINGS WITH UI BATTERIES SHIPPING82 lbs 142 lbs 100 lbs DRIVE WHEELS/TIRES (FLAT FREE)12½ inches x 2¼ inches CASTERS W/PRECISION SEALED BEARINGS8 x 1¾ inches semi-pneumatic ANTI-TIPPERSRemovable FOOTRESTS/LEGRESTSSwingaway, Removable ARMRESTSAdjustable Height - Desk UPHOLSTERYBlack Nylon BATTERIES (NOT SUPPLIED)Use only U1 batteries (Quantity - 2 U1) SPEED (M.P.H.)0 to 4 TURNING RADIUS CHAIR ONLY CHAIR WITH FOOTREST CHAIR WITH ANTI-TIPPERSFront - 26 inches; Rear - 13 inches 36 inches 16 inches *RANGE (VARIABLE)12-14 miles with U1 batteries **WEIGHT LIMITATION250 lbs

*NOTE: Values for range are calculated for maximum chair weight rating using largest batteries applicable (U1), per test procedures described in ANSI/RESNA WC/VOL2-1998 Section 4 and meet federal reimbursement requirements for this product. While considered typical, they are derived based on certain ideal conditions. Variances in battery condition, user weight, usage pattern or overall terrain conditions will result in actual values for range that differ from these stated values. Users should become accustomed to how their unique conditions impact their individual results. Users should become familiar with the battery discharge indicator on the joystick to determine the range of their wheelchair. Refer to When to Charge Batteries on page 46 for more information about the battery discharge indicator.

**NOTE: Includes seating system and accessories.

Page 9 of the Invacare XP6500 user manual — DO NOT OPERATE WITHOUT
Page 9 · click to enlarge
Page 9

LABEL LOCATIONS

LABEL LOCATIONS

! " #$ % $ & ' ( %

Refer to Owner's Manual for proper anti-tipper setting. 1085379

) !* + , *!-+ .

! / $ % $ $ !* + $$ , *!-+ . $ & ( $ 01"2#1

DO NOT OPERATE WITHOUT

THE ANTI-TIP TUBES

INSTALLED. REV. 5/98 P/N 60106X144

Show 7 more labels / short text on this page
Crossmember · Lower Frame · Tube · · · ·
Page 10 of the Invacare XP6500 user manual — Operating Information
Page 10 · click to enlarge
Page 10

SECTION 1—GENERAL GUIDELINES

SECTION 1—GENERAL GUIDELINES

WARNING This section contains important information for the safe operation and use of this product. DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as Owner’s Manuals, Service Manuals or Instruction Sheets supplied with this product or optional equipment. If you are unable to understand the Warnings, Cautions or Instructions, contact a healthcare professional, dealer or technical personnel before attempting to use this equipment - otherwise, injury or damage may occur.

Controller Settings/Repair or Service

Wheelchairs should be examined during maintenance for signs of corrosion (water exposure, incontinence, etc.). Electrical components damaged by corrosion should be replaced IMMEDIATELY. Wheelchairs that are used by incontinent users and/or are frequently exposed to water may require replacement of electrical components more frequently. Set-up of Electronics Control Unit is to be performed only by a qualified technician. The final adjustments of the controller may affect other activities of the wheelchair. Damage to the equipment could occur if improperly set-up or adjusted.

Operating Information

Performance adjustments should only be made by professionals of the healthcare field or persons fully conversant with this process and the driver's capabilities. Incorrect settings could cause injury to the driver, bystanders, damage to the chair and to surrounding property.

After the wheelchair has been set-up, check to make sure that the wheelchair performs to the specifications entered during the set-up procedure. If the wheelchair does not perform to specifications, turn the wheelchair off immediately and reenter set-up specifications. Repeat this procedure until the wheelchair performs to specifications.

ALWAYS shift your weight in the direction you are turning. DO NOT shift your weight in the opposite direction of the turn. Shifting your weight in the opposite direction of the turn may cause the inside drive wheel to lose traction and the wheelchair to tip over.

DO NOT shift your weight or sitting position toward the direction you are reaching as the wheelchair may tip over.

Page 11 of the Invacare XP6500 user manual — SECTION 1—GENERAL GUIDELINES
Page 11 · click to enlarge
Page 11

SECTION 1—GENERAL GUIDELINES

DO determine and establish your particular safety limits by practicing bending, reaching and transferring activities in the presence of a qualified healthcare professional before attempting active use of the wheelchair.

DO NOT attempt to reach objects if you have to move forward in the seat.

DO NOT attempt to reach objects if you have to pick them up from the floor by reaching down between your knees.

DO NOT lean over the top of the back upholstery to reach objects from behind as this may cause the wheelchair to tip over.

DO NOT use an escalator to move a wheelchair between floors. Serious bodily injury may occur.

Before attempting to transfer in or out of the wheelchair, every precaution should be taken to reduce the gap distance. Turn both casters parallel to the object you are transferring onto. Also, be certain the power is off and the wheel locks are engaged to prevent the wheels from moving.

DO NOT engage or disengage the motor release levers until the power is in the off position.

ProBasics strongly recommends proceeding down ramps or slopes at half speed or slower and to avoid hard braking or sudden stops.

DO NOT attempt to lift the wheelchair by lifting on any removable (detachable) parts. Lifting by means of any removable (detachable) parts of a wheelchair may result in injury to the user or damage to the wheelchair.

Anti-tippers MUST be used at all times. When outdoors on wet, soft ground or gravel surfaces, anti-tippers may not provide the same level of protection against tipover. Extra caution must be observed when traversing such surfaces.

Wheel locks are not brakes. DO NOT attempt to stop a moving wheelchair with the wheel locks.

DO NOT engage or disengage the clutches until the power is in the Off position.

DO NOT operate on roads, streets or highways.

DO NOT climb, go up or down ramps or traverse slopes greater than 9°.

DO NOT attempt to move up or down an incline with a water, ice or oil film.

NEVER leave an unoccupied wheelchair on an incline.

DO NOT attempt to drive over curbs or obstacles. Doing so may cause your wheelchair to turn over and cause bodily harm or damage to the chair.

DO NOT leave the power button in the on position when entering or exiting your wheelchair.

DO NOT stand on the frame of the wheelchair.

DO NOT use the footplates as a platform. When getting in or out of the wheelchair, make sure that the footplates are in the upward position or swing the footrests toward the outside of the chair.

Page 12 of the Invacare XP6500 user manual — Batteries
Page 12 · click to enlarge
Page 12

SECTION 1—GENERAL GUIDELINES

DO NOT stand on the front riggings, otherwise damage may occur. When getting in or out of the wheelchair, make sure that the footplates on the front riggings are in the upward position or moved out of the way.

ALWAYS wear your seat positioning strap. The seat positioning strap is a positioning belt only. It is not designed for use as a safety device withstanding high stress loads such as auto or aircraft safety belts. If signs of wear appear, belt must be replaced immediately.

Always verify that hand grips on the rear cane are secure prior to use when an assistant is used to propel or lift the chair. Check for any signs of looseness or deterioration and if found, contact a qualified technician. Do not attempt to move the wheelchair by using the hand grips if they are found to be unsecure or have deteriorated.

When cleaning rear cane or hand grip areas use only a clean towel lightly dampened with cool water. Verify that grips are dry prior to use. Use of soap or ammonia based cleaning solutions will result in the hand grips sliding off the cane assembly. Failure to observe this warning may result in injury to the user or bystanders.

If wheelchair is exposed to extreme temperature (above 100°F or Below 32°F), high humidity and/or becomes wet, prior to use, ensure handgrips do not twist on wheelchair handle - otherwise damage or injury may occur.

DO NOT sit or transfer into the wheelchair unless it is fully open and the seat frame rails are fully seated into the side frame H-blocks.

Before performing any maintenance, adjustment or service verify that On/Off switch on the joystick is in the Off position.

Avoid storage or use near external flame or combustible products.

Tire Pressure

DO NOT use your wheelchair unless it has the proper tire pressure (P.S.I.).

DO NOT overinflate the tires. Failure to follow these suggestions can cause the tire to explode and cause bodily harm. The recommended tire pressure is listed on the side wall of the tire.

Electrical Accessories

EXTREME care should be exercised when using oxygen in close proximity to electric circuits and other combustible materials. Contact your oxygen supplier for instruction in the use of oxygen.

Batteries

The warranty and performance specifications contained in this manual are based on the use of deep cycle gel cell or sealed lead acid batteries. ProBasics strongly recommends their use as the power source for this unit.

Page 13 of the Invacare XP6500 user manual — SECTION 1—GENERAL GUIDELINES
Page 13 · click to enlarge
Page 13

SECTION 1—GENERAL GUIDELINES

If wet cell batteries are used, follow all manufacturer guidelines and instructions for installation and maintenance. Regardless of battery type, proper battery footprint and use of battery terminal caps as defined in Batteries on page 46 must be followed.

Carefully read battery/battery charger information prior to installing, servicing or operating your wheelchair.

NEVER attempt to recharge the batteries by attaching cables directly to the battery terminals.

DO NOT attempt to recharge the batteries and operate the wheelchair at the same time.

DO NOT operate wheelchair with extension cord attached to the AC cable.

DO NOT attempt to recharge the batteries when the wheelchair has been exposed to ANY type of moisture.

DO NOT attempt to recharge the batteries when the wheelchair is outside.

DO NOT sit in the wheelchair while charging the batteries.

DO NOT attempt to recharge batteries using at the same time BOTH the on-board battery charger AND an independent battery charger plugged into the joystick charger port. Doing so will reduce the life of the batteries.

READ and carefully follow the manufacturer’s instructions for each charger (supplied or purchased). If charging instructions are not supplied, consult a qualified technician for proper procedures.

After charging batteries, ALWAYS make sure that the battery charger cord is unplugged. Failure to do this could result in damage to the cord or personal injury to the user or bystanders.

When using an extension cord, use only a three wire extension cord having at least 16 AWG (American Wire Gauge) wire and the same or higher electrical rating as the device being connected.

Use of improper extension cord could result in risk of fire and electric shock.

Ensure the pins of the extension cord plug are the same number, size, and shape as those on the charger.

DO NOT under any circumstances cut or remove the round grounding plug from the charger AC cable plug or the extension cord plug.

Three prong to two prong adapters should not be used. Use of three prong adapters can result in improper grounding and present a shock hazard to the user.

Charging Batteries

Page 14 of the Invacare XP6500 user manual — Rain Test
Page 14 · click to enlarge
Page 14

SECTION 1—GENERAL GUIDELINES

Grounding Instructions

DANGER When using an extension cord having at least 16 AWG (American Wire Gauge) wire and the same or higher electrical rating as the device being connected. Use of improper extension cord could result in a risk of fire and electric shock. Three prong adapters can result in improper grounding and present a shock hazard to the user.

DO NOT, under any circumstances, cut or remove the round grounding prong from any plug used with or for ProBasics products. Some devices are equipped with three-prong (grounding) plugs for protection against possible shock hazards and fire. Where a two-prong wall receptacle is encountered, it is the personal responsibility and obligation of the customer to contact a qualified electrician and have the two-prong receptacle replaced with a properly grounded three-prong wall receptacle in accordance with the National Electrical Code. If you must use an extension cord, use ONLY a three-wire extension cord having the same or higher electrical rating as the device being connected. In addition, ProBasics has placed RED/ORANGE warning tags on some equipment. DO NOT remove these tags.

Rain Test

ProBasics has tested its power wheelchairs in accordance with ISO 7176 “Rain Test.” This provides the end user or his/her assistant sufficient time to remove his/her power wheelchair from a rain storm and retain wheelchair operation.

DO NOT leave power wheelchair in a rain storm of any kind.

DO NOT use power wheelchair in a shower or leave it in a damp bathroom while taking a shower.

DO NOT store power wheelchair in a damp area for an extended period of time.

Direct exposure to excessive rain or dampness may cause the chair to malfunction electrically and mechanically, may cause the chair to prematurely rust or may damage the upholstery.

Check to ensure that the RED and GREY battery terminal caps are secured in place, joystick boot is NOT torn or cracked where water can enter and that all electrical connections are secure at all times.

DO NOT use the wheelchair if the joystick boot is torn or cracked. If the joystick boot becomes torn or cracked, replace IMMEDIATELY.

Page 15 of the Invacare XP6500 user manual — Weight Limitation
Page 15 · click to enlarge
Page 15

SECTION 1—GENERAL GUIDELINES

Weight Training

ProBasics DOES NOT recommend the use of its wheelchairs as a weight training apparatus. ProBasics wheelchairs have not been designed or tested as a seat for any kind of weight training. If occupant uses said wheelchair as a weight training apparatus, ProBasics shall NOT be liable for bodily injury and the warranty is void.

Weight Limitation

The wheelchair has a weight limitation, including seating system and accessories, of 250 lbs. Refer to Percentage of Weight Distribution on page 22. ProBasics recommends that only heavy-duty wheelchairs should be used for individuals weighing more than 250 lbs. Further, the activity level of the individual wheelchair user is important. For instance, a 170 lb. active wheelchair could subject the wheelchair to more stress than a 250 lb. user. ProBasics recommends that very active users consider the use of heavy-duty wheelchairs.

Page 16 of the Invacare XP6500 user manual — SECTION 2—EMI INFORMATION
Page 16 · click to enlarge
Page 16

SECTION 2—EMI INFORMATION

SECTION 2—EMI INFORMATION

CAUTION: IT IS VERY IMPORTANT THAT YOU READ THIS INFORMATION REGARDING THE POSSIBLE EFFECTS OF ELECTROMAGNETIC INTERFERENCE ON YOUR POWERED WHEELCHAIR. Electromagnetic Interference (EMI) From Radio Wave Sources Powered wheelchairs and motorized scooters (in this text, both will be referred to as powered wheelchairs) may be susceptible to electromagnetic interference (EMI), which is interfering electromagnetic energy (EM) emitted from sources such as radio stations, TV stations, amateur radio (HAM) transmitters, two way radios, and cellular phones. The interference (from radio wave sources) can cause the powered wheelchair to release its brakes, move by itself, or move in unintended directions. It can also permanently damage the powered wheelchair's control system. The intensity of the interfering EM energy can be measured in volts per meter (V/m). Each powered wheelchair can resist EMI up to a certain intensity. This is called its "immunity level." The higher the immunity level, the greater the protection. At this time, current technology is capable of achieving at least a 20 V/m immunity level, which would provide useful protection from the more common sources of radiated EMI. There are a number of sources of relatively intense electromagnetic fields in the everyday environment. Some of these sources are obvious and easy to avoid. Others are not apparent and exposure is unavoidable. However, we believe that by following the warnings listed below, your risk to EMI will be minimized. The sources of radiated EMI can be broadly classified into three types: 1) Hand-held Portable transceivers (transmitters-receivers with the antenna mounted directly on the transmitting unit. Examples include: citizens band (CB) radios, "walkie talkie", security, fire and police transceivers, cellular telephones, and other personal communication devices). NOTE: Some cellular telephones and similar devices transmit signals while they are ON, even when not being used. 2) Medium-range mobile transceivers, such as those used in police cars, fire trucks, ambulances and taxis. These usually have the antenna mounted on the outside of the vehicle; and 3) Long-range transmitters and transceivers, such as commercial broadcast transmitters (radio and TV broadcast antenna towers) and amateur (HAM) radios. NOTE: Other types of hand-held devices, such as cordless phones, laptop computers, AM/FM radios, TV sets, CD players, cassette players, and small appliances, such as electric shavers and hair dryers, so far as we know, are not likely to cause EMI problems to your powered wheelchair.

Page 17 of the Invacare XP6500 user manual — SECTION 2—EMI INFORMATION
Page 17 · click to enlarge
Page 17

SECTION 2—EMI INFORMATION

Powered Wheelchair Electromagnetic Interference (EMI) Because EM energy rapidly becomes more intense as one moves closer to the transmitting antenna (source), the EM fields from hand-held radio wave sources (transceivers) are of special concern. It is possible to unintentionally bring high levels of EM energy very close to the powered wheelchair's control system while using these devices. This can affect powered wheelchair movement and braking. Therefore, the warnings listed below are recommended to prevent possible interference with the control system of the powered wheelchair. Electromagnetic interference (EMI) from sources such as radio and TV stations, amateur radio (HAM) transmitters, two-way radios, and cellular phones can affect powered wheelchairs and motorized scooters. FOLLOWING THE WARNINGS LISTED BELOW SHOULD REDUCE THE CHANCE OF UNINTENDED BRAKE RELEASE OR POWERED WHEELCHAIR MOVEMENT WHICH COULD RESULT IN SERIOUS INJURY. 1) Do not operate hand-held transceivers (transmitters receivers), such as citizens band (CB) radios, or turn ON personal communication devices, such as cellular phones, while the powered wheelchair is turned ON; 2) Be aware of nearby transmitters, such as radio or TV stations, and try to avoid coming close to them; 3) If unintended movement or brake release occurs, turn the powered wheelchair Off as soon as it is safe; 4) Be aware that adding accessories or components, or modifying the powered wheelchair, may make it more susceptible to EMI. NOTE: There is no easy way to evaluate their effect on the overall immunity of the powered wheelchair). 5) Report all incidents of unintended movement or brake release to the powered wheelchair manufacturer, and note whether there is a source of EMI nearby. Important Information

1. 20 volts per metre (V/m) is a generally achievable and useful immunity level against EMI (as of May 1994) (the higher the level, the greater the protection); 2) This device has been tested to a radiated immunity level of 20 volts per meter. 3) The immunity level of the product is unknown. Modification of any kind to the electronics of this wheelchair as manufactured by Invacare may adversely affect the EMI immunity levels.
Page 18 of the Invacare XP6500 user manual — Coping With Everyday Obstacles
Page 18 · click to enlarge
Page 18

SECTION 3—SAFETY/HANDLING OF WHEELCHAIRS

SECTION 3—SAFETY/HANDLING OF WHEELCHAIRS

“Safety and Handling” of the wheelchair requires the close attention of the wheelchair user as well as the assistant. This manual points out the most common procedures and techniques involved in the safe operation and maintenance of the wheelchair. It is important to practice and master these safe techniques until you are comfortable in maneuvering around the frequently encountered architectural barriers.

Use this information only as a “basic” guide. The techniques that are discussed on the following pages have been used successfully by many.

Individual wheelchair users often develop skills to deal with daily living activities that may differ from those described in this manual. ProBasics recognizes and encourages each individual to try what works best for him/her in overcoming architectural obstacles that they may encounter. However, all warnings and cautions given in this manual MUST be followed. Techniques in this manual are a starting point for the new wheelchair user and assistant with “safety” as the most important consideration for all.

Stability and Balance

WARNING ALWAYS wear your seat positioning strap.The seat positioning strap is a positioning belt only. It is not designed for use as a safety device withstanding high stress loads such as auto or aircraft safety belts. If signs of wear appear, belt MUST be replaced immediately.

To assure stability and proper operation of your wheelchair, you must at all times maintain proper balance. Your wheelchair has been designed to remain upright and stable during normal daily activities as long as you do not move beyond the center of gravity. DO NOT lean forward out of the wheelchair any further than the length of the armrests.

Coping With Everyday Obstacles

Coping with the irritation of everyday obstacles can be alleviated somewhat by learning how to manage your wheelchair. Keep in mind your center of gravity to maintain stability and balance.

While the walking beam allows you to traverse up to a 2-inch bump or threshold, stopping after the wheels cross the bump poses a problem. The chair cannot reverse over the bump at this point. Continue forward and then turn around.

While the wheelchair is designed for use primarily in and around the home, the provider should determine whether this chair is suitable for the actual environment the chair will be used in.

Page 19 of the Invacare XP6500 user manual — Tipping
Page 19 · click to enlarge
Page 19

SECTION 3—SAFETY/HANDLING OF WHEELCHAIRS

DO NOT go down ramp at full speed. Some seat/back positions will cause wheelchair to feel unstable.

A Note to Wheelchair Assistants

When assistance to the wheelchair user is required, remember to use good body mechanics. Keep your back straight and bend your knees whenever tilting wheelchair or traversing curbs or other impediments.

Also, be aware of detachable parts such as arms or legrests. These must NEVER be used to move the wheelchair or as lifting supports, as they may be inadvertently released, resulting in possible injury to the user and/or assistant(s).

When learning a new assistance technique, have an experienced assistant help you before attempting it alone.

Tipping

WARNING DO NOT tip the wheelchair without assistance.

When tipping the wheelchair, an assistant should grasp the back of the wheelchair on a non-removable (non-detachable) part. Inform the wheelchair occupant before tipping the wheelchair and remind him/her to lean back. Be sure the occupant’s feet and hands are clear of all wheels.

After mastering the techniques of tipping the wheelchair, use this procedure to tackle curbs, short stairs, etc.

This procedure requires two assistants. The second assistant should be positioned at the front of the wheelchair lifting upward on a non-removable (non-detachable) part of the wheelchair frame when lifting the wheelchair and stabilizing the wheelchair when the wheelchair is being lowered to the ground.

The first assistant should turn the anti-tippers so the wheels are pointing up, apply a continuous downward motion until the balance point is achieved and the front casters clear the curb. At this point, the assistants will feel a difference in the weight distribution.

Page 20 of the Invacare XP6500 user manual — Lifting/Stairways
Page 20 · click to enlarge
Page 20

SECTION 3—SAFETY/HANDLING OF WHEELCHAIRS

WARNING DO NOT let the wheelchair drop the last few inches to the ground. This could result in injury to the occupant.

Roll the wheelchair forward and SLOWLY lower the wheelchair in one continuous movement. Push the wheelchair forward until the rear wheels roll up and over the curb.

WARNING Make sure anti-tipper wheels are pointing down towards the ground/floor before using the wheelchair.

Turn the anti-tipper wheels down towards the ground/floor.

Lifting/Stairways

DO NOT attempt to move an occupied power wheelchair between floors using a stairway. Use an elevator to move an occupied power wheelchair between floors. If moving a power wheelchair between floors by means of a stairway, the occupant MUST be removed and transported independently of the power wheelchair. Extreme caution is advised when it is necessary to move an unoccupied power wheelchair up or down the stairs. ProBasics recommends using two assistants and making thorough preparations. Use ONLY secure, nondetachable parts for hand-hold supports. It is strongly recommended to lift the wheelchair only by the rear frame and the front forks - otherwise injury or damage may occur. DO NOT attempt to lift the wheelchair by any removable (detachable) parts. Lifting by means of any removable (detachable) parts of a wheelchair may result in injury to the user or damage to the wheelchair. The weight of the wheelchair without the user and batteries is between 80 and 127 lbs.

Follow this procedure for moving the wheelchair between floors when an elevator is NOT available or lifting the wheelchair is necessary: NOTE: When using a stairway to move the wheelchair, seat and any accessories, move all wheelchair components away from the stairway prior to reassembly.

1. Remove the occupant from the wheelchair.
2. Remove the battery boxes from the wheelchair. Refer to Removing/Installing the Battery Boxes on page 49.
3. If necessary, fold the wheelchair. Refer to Transporting the Wheelchair on page 61.
4. Bend your knees and keep your back straight.
Page 21 of the Invacare XP6500 user manual — Transferring to and From Other Seats
Page 21 · click to enlarge
Page 21

SECTION 3—SAFETY/HANDLING OF WHEELCHAIRS

5. Using non-removable (non-detachable) parts of the wheelchair, lift the wheelchair off the ground and transfer the wheelchair up or down the stairs.
6. The wheelchair should not be lowered until the last stair has been negotiated and the wheelchair has been carried away from the stairway.

ESCALATORS WARNING DO NOT use an escalator to move a wheelchair between floors. Serious bodily injury may occur.

Transferring to and From Other Seats

ALWAYS turn the wheelchair power Off and engage the clutches to prevent the wheels from moving before attempting to transfer in or out of the wheelchair. Also make sure every precaution is taken to reduce the gap distance by turning both casters parallel to the object you are transferring onto.

Invacare XP6500 — Transferring to and From Other Seats diagram (page 21)
Fig. 2 · Transferring to and From Other Seats
1. Position the wheelchair as close as possible along side the seat to which you are transferring, with the casters pointing parallel to it.
2. After the wheelchair is positioned properly for transfer, verify that the clutches are engaged. Refer to Engaging/Disengaging the Clutches on page 56.

FIGURE 3.1 Transferring to and From

3. Shift body weight into seat with transfer.

During independent transfer, little or no seat platform will be beneath you. Use a transfer board if at all possible.

Minimum Gap Distance · Other Seats

Page 22 of the Invacare XP6500 user manual — Reaching, Leaning and Bending - Forward
Page 22 · click to enlarge
Page 22

SECTION 3—SAFETY/HANDLING OF WHEELCHAIRS

Percentage of Weight Distribution

DO NOT attempt to reach objects if you have to move forward in the seat or pick them up from the floor by reaching down between your knees. Proper positioning is essential for your safety. When reaching, leaning, bending or bending forward, it is important to use the casters as a tool to maintain stability and balance. Many activities require the wheelchair user to reach, bend and transfer in and out of the wheelchair. These movements will cause a change to the normal balance, center of gravity, and weight distribution of the wheelchair. To determine and establish your particular safety limits, practice bending, reaching and transferring activities in several combinations in the presence of a qualified healthcare professional before attempting active use of the wheelchair.

Reaching, Leaning and Bending - Forward

1. Position the front casters as far forward as possible.
2. Engage clutches and, if applicable, wheel locks*.

DO NOT attempt to reach objects if you have to move forward in the seat or pick them up from the floor by reaching down between your knees. *Wheel locks are an option. You can order the wheelchair with or without wheel locks. Transfer to and from the wheelchair in the presence of a qualified healthcare professional to determine individual safety limits. ProBasics strongly recommends ordering the wheel locks as an additional safeguard for the wheelchair user.

Page 23 of the Invacare XP6500 user manual — Reaching and Bending - Backward
Page 23 · click to enlarge
Page 23

SECTION 3—SAFETY/HANDLING OF WHEELCHAIRS

Invacare XP6500 — SECTION 3—SAFETY/HANDLING OF WHEELCHAIRS diagram (page 23)
Fig. 3
Invacare XP6500 — SECTION 3—SAFETY/HANDLING OF WHEELCHAIRS diagram (page 23)
Fig. 4

FIGURE 3.2 Reaching, Leaning and Bending - Forward

Reaching and Bending - Backward

WARNING DO NOT lean over the top of the back upholstery. This will change your center of gravity and may cause you to tip over.

1. Position wheelchair as close as possible to the desired object.
2. Point the front and rear casters rearward to create the longest possible wheelbase.
3. Reach back only as far as your arm will extend without changing your sitting position.
Invacare XP6500 — Reaching and Bending - Backward diagram (page 23)
Fig. 5 · Reaching and Bending - Backward

FIGURE 3.3 Reaching and Bending - Backward

Page 24 of the Invacare XP6500 user manual — Safety Inspection Checklists
Page 24 · click to enlarge
Page 24

SECTION 4—SAFETY INSPECTION/TROUBLESHOOTING

SECTION 4—SAFETY INSPECTION/TROUBLESHOOTING

NOTE: Every six months or as necessary take your wheelchair to a qualified dealer for a thorough inspection and servicing. Regular cleaning will reveal loose or worn parts and enhance the smooth operation of your wheelchair. To operate properly and safely, your wheelchair must be cared for just like any other vehicle. Routine maintenance will extend the life and efficiency of your wheelchair.

Safety Inspection Checklists

Initial adjustments should be made to suit your personal body structure needs and preference. Thereafter follow these maintenance procedures:

❑Ensure that the wheelchair rolls straight (no excessive drag or pull to one side).

❑Ensure that all fasteners on the clothing guards are secure.

❑Ensure that the arms are secure but easy to release and adjustment levers engage

properly.

❑Ensure that adjustable height arms operate and lock securely.

❑Ensure armrest pads sit flush against arm tubes.

❑Ensure seat is secured to wheelchair frame.

❑Clean seat upholstery and armrests.

❑Ensure wheel mounting nuts are secure on drive wheels.

❑Ensure that casters are free of debris.

❑Ensure no excessive side movement or binding occurs when drive wheels are lifted

and spun when disengaged (free-wheeling).

❑Ensure that drive wheel axle bolts and locking tab washers are secure.

❑Ensure that the wheel locks engage properly and do not interfere with the tires when

rolling. Also, ensure that the pivot points are free of wear and looseness.

❑Inspect caster assembly has proper tension when caster is spun. Caster should come

to a gradual stop.

❑Loosen/tighten caster locknut if wheel wobbles noticeably or binds to a stop.

❑Ensure all caster/wheel/fork/headtube fasteners are secure.

Inspect/Adjust Initially

Page 25 of the Invacare XP6500 user manual — SECTION 4—SAFETY INSPECTION/TROUBLESHOOTING
Page 25 · click to enlarge
Page 25

SECTION 4—SAFETY INSPECTION/TROUBLESHOOTING

❑Inspect tires for flat spots and wear. If tires are pneumatic, check for proper inflation.

❑Seat and/or back upholstery have no rips and do not sag. Replace if necessary.

❑Seat is secured to wheelchair frame.

❑Seat release latch is not worn and is functional. Replace if necessary.

❑Inspect tires for flat spots and wear. If the tires are pneumatic, check for proper

inflation.

❑Ensure all caster/wheel/fork/headtube fasteners are secure.

❑Ensure that casters are free of debris.

❑Ensure arm pivot points are not worn and/or loose. Replace if necessary.

❑Ensure wheel mounting nuts are secure on drive wheels.

❑Ensure no excessive side movement or binding occurs when drive wheels are lifted

and spun when disengaged (free-wheeling).

❑Ensure that drive wheel axle bolts and locking tab washers are secure.

❑Ensure that casters are free of debris.

❑Inspect caster assembly has proper tension when caster is spun. Caster should come

to a gradual stop.

❑Loosen/tighten caster locknut if wheel wobbles noticeably or binds to a stop.

❑Ensure that the wheel locks engage properly and do not interfere with the tires when

rolling. Also, ensure that the pivot points are free of wear and looseness.

❑Inspect the seat positioning strap for signs of wear. Replace if worn or damaged.

❑Ensure that the buckle on the seat positioning strap latches. Replace if necessary.

❑Verify that the hardware that attaches the seat positioning strap to the seat frame is

secure and undamaged. Replace if necessary.

❑Ensure wheelchair rolls straight (no excessive drag or pull to one side).

❑Ensure that casters are free of debris.

❑Ensure that all fasteners on the clothing guards are secure.

❑Ensure arms are secure but easy to release and adjustment levers engage properly.

❑Ensure adjustable height arms operate and lock securely.

❑Ensure arm pivot points are not worn and/or loose. Replace if necessary.

❑Ensure armrest pads sit flush against arm tubes.

Inspect/Adjust Weekly · Inspect/Adjust Monthly · Inspect/Adjust Periodically

Page 26 of the Invacare XP6500 user manual — SYMPTOM PROBABLE CAUSE SOLUTIONS
Page 26 · click to enlarge
Page 26

SECTION 4—SAFETY INSPECTION/TROUBLESHOOTING

❑Ensure seat and/or back upholstery have no rips and do not sag. Replace if necessary.

❑Ensure seat release latch is not worn. Replace if necessary.

❑Seat and/or back upholstery have no rips and do not sag. Replace if necessary.

❑Clean upholstery and armrests.

❑Inspect charger AC power cord for damage. Replace if necessary.

❑Check that all labels are present and legible. Replace if necessary.

Troubleshooting - Mechanical

If pneumatic, check tires for correct and equal pressure.

X

Check for loose stem nuts/bolts.

X

Check that both casters contact ground at the same time.

Troubleshooting Guide

SYMPTOM PROBABLE CAUSE SOLUTIONS

Batteries draw excessive current when charging.

Battery failure.

Have batteries checked for shorted cell. Replace if necessary. Contact dealer for service.

Electrical malfunction.

Battery indicator flashes the charge level is low immedi- ately after recharging.

Battery failure.

Check batteries for shorted cell. Replace if nec- essary. Contact dealer/ProBasics. Poor connections between charger and wheel- chair. Contact dealer for service.

Charger malfunction. Electrical malfunction.

Battery indicator flashes the charge level is low too soon after being recharged.

Batteries not charged. Weak batteries. Electrical malfunction.

Have charger checked. Replace batteries if necessary. Contact dealer for service.

Motor chatters or runs irregularly.

Electrical malfunction. Contact dealer.

Show 7 more labels / short text on this page
Chair Veers Left/Right · Sluggish Turn/Performance · Casters Flutter · Squeaks and Rattles · Looseness In Chair · Chair 3 Wheels · Solutions
Page 27 of the Invacare XP6500 user manual — DON’T DO
Page 27 · click to enlarge
Page 27

SECTION 4—SAFETY INSPECTION/TROUBLESHOOTING

SYMPTOM PROBABLE CAUSE SOLUTIONS

Wheelchair does not respond to commands. Power on, battery indicator flashes.

One or both clutches disen- gaged.

Engage motor locks/clutches.

Only one drive wheel turns. Electrical malfunction. One clutch is disengaged.

Contact dealer for service. Engage clutch.

Joystick erratic or does not respond as desired.

Electrical malfunction. Controller programmed improperly.

Contact dealer for service. Reprogram controller. Contact dealer for ser- vice.

Wheelchair does not respond to commands. Power indicator off - even after recharging.

Poor battery terminal con- nection. Electrical malfunction.

Clean terminals.

Contact dealer.

Checking Battery Charge Level

DON’T DO

Don’t perform any installation or maintenance with- out first reading this manual.

Read and understand this manual and any service infor- mation that accompanies a battery and charger before operating the wheelchair.

Don’t make it a habit to discharge batteries to the lowest level.

Recharge as frequently as possible to maintain a high charge level and extend battery life.

Don’t use randomly chosen batteries/chargers.

Follow recommendations in this manual when selecting a battery or charger.

Don’t put new batteries into service before charging.

Fully charge a new battery before using. Use a carrying strap to remove, move or install a battery.

Don’t tip or tilt batteries.

Don’t tap on clamps or terminals with tools.

Push battery clamps onto terminals. Spread clamps wider if necessary.

Don’t mismatch your battery and chargers. Don’t use randomly chosen batteries or chargers.

Use ONLY a GEL charger for a GEL or sealed battery and a regular charger for regular batteries.

Page 28 of the Invacare XP6500 user manual — Joystick Switches and Indicators
Page 28 · click to enlarge
Page 28

SECTION 5—WHEELCHAIR OPERATION

SECTION 5—WHEELCHAIR OPERATION

After ANY adjustments, repair or service and before use, make sure that all attaching hardware is tightened securely - otherwise injury or damage may result. Set-up of Electronics Control Unit is to be performed only by a qualified technician. The final adjustments of the controller may affect other activities of the wheelchair. Damage to the equipment could occur if improperly set-up or adjusted.

Joystick Switches and Indicators

Multi Function Charger Port

The multi function charger port is located at the front of the joystick housing. This provides easy access for charging the wheelchair batteries. This port also serves as the Remote Programmer Communication connection.

This toggle switch is located at the back of the joystick housing.

The speed control switch is located on the back of the joystick housing. This rotary switch is used for controlling the speed and acceleration of the wheelchair.

1. Turn the switch clockwise to increase the speed of the wheelchair.
2. Turn the switch counterclockwise to decrease the speed of the wheelchair.

The joystick has proportional drive control, meaning that further the joystick is pushed from the upright (neutral) position, the faster the wheelchair moves. Your top speed, however, is limited by the setting of the speed-control knob and programmed settings.

To slow the wheelchair to a stop, simply release the joystick. The wheelchair has automatic speed and direction compensation to minimize corrections.

On/Off Toggle Switch · Speed Control switch · Joystick

Page 29 of the Invacare XP6500 user manual — SECTION 5—WHEELCHAIR OPERATION
Page 29 · click to enlarge
Page 29

SECTION 5—WHEELCHAIR OPERATION

FIGURE 5.1 Joystick Switches and Indicators

The Information Gauge Display is located on the front of the joystick housing. It provides the following information to the user on the status of the wheelchair -

1. Power is on.
2. True state-of-battery-charge, including notification of when the battery requires charging:

A. GREEN LED is lit, indicating well charged batteries.

B. Only AMBER LEDs are lit, indicating batteries are moderately charged. Recharge

batteries before taking a long trip.

C. Only RED LED is lit, indicating batteries are running out of charge. Recharge

batteries as soon as possible.

3. Program, inhibit or charge modes.
4. Fault indication (Flash Codes).

The Information Gauge display also serves as a system diagnostic device when a fault is detected by the control module. A specific number of flashes of the LEDs indicate the type of fault detected. Refer to the following table of the diagnostic indications of the wheelchair status.

Show 10 more labels / short text on this page
Joystick · Horn · Information · Gauge On/Off · Button · Speed Control Knob · To Controller · Multi-Function · Charger Port · Information Gauge Display
Page 30 of the Invacare XP6500 user manual — SECTION 5—WHEELCHAIR OPERATION
Page 30 · click to enlarge
Page 30

SECTION 5—WHEELCHAIR OPERATION

Invacare XP6500 — SECTION 5—WHEELCHAIR OPERATION diagram (page 30)
Fig. 6

All LEDs are on. Power is On. Fewer than three LEDs on implies reduced battery charge.

Left RED LED is flashing. Battery charge is low. The batteries should be charged as soon as possi- ble.

The steady LEDs indicate the current state of the battery charge.

The current state of bat- tery charge will be dis- played at the same time.

All LEDs are flashing slowly. Joystick has detected Out-of-Neu- tral-at-Power-Up mode.

Joystick uses Flash codes to indicate faults. Refer to the electronics manual (Part Number 1123804).

Page 31 of the Invacare XP6500 user manual — SECTION 5—WHEELCHAIR OPERATION
Page 31 · click to enlarge
Page 31

SECTION 5—WHEELCHAIR OPERATION

Operating the Wheelchair

Turning the Power On/Off

1. To turn the power On, press the On/Off button.

A. The Current Battery Charge - Shows all LEDs lit or partial LEDs lit. Refer to the

preceding information gauge display table.

B. Out Of Neutral At Power up - Shows all LEDs flashing slowly. This occurs when

the power is turned on when the joystick is out of neutral. This feature prevents sudden and unexpected movements of the power chair.

2. To turn the power off, press the On/Off button.
1. Press the horn button located above the information gauge on the joystick housing.

Using the Joystick to Drive the Chair

The joystick is located at the front of the joystick housing and provides smooth control of speed and direction. It is equipped with 360 degrees of mobility for ease of operation. The joystick is spring-loaded, and automatically returns to the upright (neutral) position when released. Pushing the joystick in a given direction causes the chair to move in that direction.

The joystick has proportional drive control, meaning that the further it is pushed from the upright (neutral) position, the faster the wheelchair moves. The maximum speed, however, is limited by the setting of the speed-control knob.

To slow the wheelchair to a stop, simply release the joystick. The wheelchair has automatic speed and direction compensation to minimize corrections.

When first learning to drive, select a slow speed and try to drive the wheelchair as slowly as possible by pushing the joystick slightly forward. This exercise will help you learn to utilize the full potential of the proportional control and allow you to start and stop smoothly.

To drive the wheelchair, perform the following:

1. Adjust speed control switch to the appropriate setting.
2. Turn the power On. Refer to Turning the Power On/Off on page 31.

Using the Horn

Page 32 of the Invacare XP6500 user manual — MOVEMENT ACTION
Page 32 · click to enlarge
Page 32

SECTION 5—WHEELCHAIR OPERATION

3. Move the joystick in the following manner:

MOVEMENT ACTION

FORWARD Push forward on the joystick.

REVERSE Pull back on the joystick.

Turn RIGHT Move the joystick RIGHT.

Turn LEFT Move the joystick LEFT.

STOP Release the joystick and the wheelchair will quickly slow down.

To Move Backward To Move Right

FIGURE 5.2 Using the Joystick to Drive the Chair

Show 6 more labels / short text on this page
Joystick · To Move · To Move Left · Forward · On/Off Button · Speed Control Knob
Page 33 of the Invacare XP6500 user manual — Installing/Removing/Using the Footrest/Legrest
Page 33 · click to enlarge
Page 33

SECTION 6—FRONT RIGGINGS

SECTION 6—FRONT RIGGINGS

After ANY adjustments, repair or service and before use, make sure that all attaching hardware is tightened securely - otherwise injury or damage may result.

Before performing any maintenance, adjustment or service verify that On/Off switch on the joystick is in the Off position.

Installing/Removing/Using the Footrest/Legrest

1. Turn the swingaway footrest assembly to the side (open footplate is perpendicular to wheelchair).
2. Install the hinge plates on the swingaway footrest assembly onto the hinge pins on the wheelchair frame.
3. Push the swingaway footrest assembly towards the inside of the wheelchair until it locks into place.
4. Repeat this procedure for the other footrest assembly.

FIGURE 6.1 Installing/Removing/Using the Footrest/Legrest

Show 8 more labels / short text on this page
Installing · Footrest Release Lever · Swingaway · Footrest Assembly · Hinge · Pins · Footplate · Hinge Plates
Page 34 of the Invacare XP6500 user manual — Adjusting the Footrest Height
Page 34 · click to enlarge
Page 34

SECTION 6—FRONT RIGGINGS

1. Push the footrest release lever inward.
2. Rotate swingaway footrest assembly outward.
3. Lift the swingaway footrest assembly off of the hinge pins.
1. To move the footrest out of the way without removing, push the footrest release lever inward, rotate footrest outward.
2. To lock the footrest, push the swingaway footrest assembly towards the inside of the wheelchair until it locks into place.

Adjusting the Footrest Height

1. Remove the swingaway footrest assembly. Refer to Installing/Removing/Using the Footrest/Legrest on page 33.
Invacare XP6500 — Adjusting the Footrest Height diagram (page 34)
Fig. 7 · Adjusting the Footrest Height
2. Remove impact guards and/or calf strap, if installed.
3. Pull the cam lock lever up to the unlocked position.
4. Push in the release buttons and reposition the lower footrest assembly to the desired height.
5. Ensure that the release buttons fully protrude from holes on both sides of the upper footrest support.
6. Rotate cam lock lever down to locked position.
8. Repeat this procedure for the other footrest, if necessary.
9. Reinstall the swingaway footrest assembly. Refer to Installing/Removing/Using the Footrest/Legrest on page 33.

FIGURE 6.2 Adjusting the Footrest Height

Removing · Using

Page 35 of the Invacare XP6500 user manual — Raising/Lowering the Legrest Assembly
Page 35 · click to enlarge
Page 35

SECTION 6—FRONT RIGGINGS

Replacing the Heel Loop

1. Pull the cam lock lever up to the unlocked position.
2. Push in the release buttons and remove the lower footrest assembly.
3. Remove the mounting screw, spacer and locknut that secure the heel loop to the footrest.
4. Remove existing heel loop from slide tube.
5. Install new heel loop onto slide tube.
6. Install the mounting screw, spacer and locknut to secure the heel loop to the footrest. Tighten until the spacer is secure.

FIGURE 6.3 Replacing the Heel Loop

7. Insert the lower footrest assembly into the upper footrest assembly to desired height.
8. Ensure that the release buttons fully protrude from holes on both sides of the upper footrest support.
9. Rotate cam lock lever down to locked position.

Raising/Lowering the Legrest Assembly

Ensure hands and fingers are clear of elevating legrest mechanism before pushing release lever to lower the elevating legrest. Otherwise injury may occur due to pinch points. The wheelchair user’s leg MUST be supported by an assistant before attempting to lower legrest.

Show 9 more labels / short text on this page
Mounting Screw · Cam Lock Lever · Spacer · Heel Loop · Slide Tube · Lower Footrest · Assembly · Footrest · Locknut
Page 36 of the Invacare XP6500 user manual — SECTION 6—FRONT RIGGINGS
Page 36 · click to enlarge
Page 36

SECTION 6—FRONT RIGGINGS

1. To raise the elevating legrest, the assistant should hold the support tube and raise elevating legrests until the desired height is obtained.
2. To lower the elevating legrest, perform the following:

A. Support user leg with one hand.

B. Push release lever downward with other hand.

C. Gently, lower user leg down and rest against the legrest.

FIGURE 6.4 Raising/Lowering the Legrest Assembly

Show 4 more labels / short text on this page
Release · Lever · Support · Tube
Page 37 of the Invacare XP6500 user manual — Adjusting Armrest Height
Page 37 · click to enlarge
Page 37

SECTION 7—ARMS

SECTION 7—ARMS

After ANY adjustments, repair or service and before use, make sure that all attaching hardware is tightened securely - otherwise injury or damage may result.

Before performing any maintenance, adjustment or service verify that On/Off switch on the joystick is in the Off position.

Adjusting Armrest Height

WARNING Make sure the height adjustment lever is in the locked position before using the wheelchair.

1. Unlock the armrest by flipping the height adjustment lever on the top front of the armrest to the up (horizontal) position.
2. Adjust armrest to one of five positions.
3. Lock the armrest by pressing the height adjustment lever into the down (vertical) position when the desired armrest height is achieved.
4. Repeat STEPS 1-3 for other armrest.

FIGURE 7.1 Adjusting Armrest Height

Show 4 more labels / short text on this page
Armrest · Locked (Vertical) · Height Adjustment Lever · Unlocked (Horizontal)
Page 38 of the Invacare XP6500 user manual — Replacing Armrest Pad
Page 38 · click to enlarge
Page 38

SECTION 7—ARMS

Swing-Back Arms

WARNING Make sure the armrest release lever is in the locked position before using the wheelchair.

1. Unlock the swing-back arms by rotating the armrest release lever towards the outside of the wheelchair.
2. Pull the front of the swing-back arm straight up/out of the arm socket and towards the rear of the wheelchair.
3. To use the swing-back arm, push the swing-back arm towards the front of the wheelchair and then downward into the arm socket.
4. Lock the swing-back arms by rotating the armrest release lever towards the inside of the wheelchair.

FIGURE 7.2 Swing-Back Arms

Replacing Armrest Pad

1. Remove the mounting screws that secure the armrest pad to the armrest assembly.
2. Replace armrest pad and securely tighten with the existing mounting screws.

FIGURE 7.3 Replacing Armrest Pad

Show 9 more labels / short text on this page
Front · Swing Back Arm · Armrest Release Lever · Rear · Arm Socket · Armrest Pad · Armrest Assembly · Mounting · Screws
Page 39 of the Invacare XP6500 user manual — SEAT DEPTH NUMBER OF SCREWS
Page 39 · click to enlarge
Page 39

SECTION 8—SEAT AND BACK

SECTION 8—SEAT AND BACK

After ANY adjustments, repair or service and before use, make sure that all attaching hardware is tightened securely - otherwise injury or damage may result.

Before performing any maintenance, adjustment or service verify that On/Off switch on the joystick is in the Off position.

Replacing the Seat Upholstery

1. Remove the mounting screws that secure the existing seat upholstery to the wheelchair frame.

SEAT DEPTH NUMBER OF SCREWS

16 inch 18 inch 8 10

2. Remove the existing seat upholstery from the wheelchair frame.
3. Install new seat upholstery by reversing Steps 1-2.

FIGURE 8.1 Replacing the Seat Upholstery

Show 4 more labels / short text on this page
Mounting Screws · 16-inch Seat Depth · Seat Upholstery · Wheelchair Frame
Page 40 of the Invacare XP6500 user manual — SECTION 8—SEAT AND BACK
Page 40 · click to enlarge
Page 40

SECTION 8—SEAT AND BACK

Replacing the Back Upholstery

1. Flip swing back arms up and out of the way. Refer to Swing-Back Arms on page 38.
2. Remove the six mounting screws and washers that secure the existing back upholstery to the back canes.
3. Cut the tie wraps that secure the existing back upholstery to the wheelchair frame.
4. Securely tighten the new back upholstery to the back canes with the six mounting screws and washers starting with the top hole of the back cane.
5. Install two new tie wraps through the bottom holes of the upholstery to the bottom of the back upholstery to the back canes.
6. Flip swing-back arms down for use. Refer to Swing-Back Arms on page 38.
Invacare XP6500 — SECTION 8—SEAT AND BACK diagram (page 40)
Fig. 8

FIGURE 8.2 Replacing the Back Upholstery

Show 6 more labels / short text on this page
Back Upholstery Washers · Mounting Screws · Back Cane · Tie Wraps · Washers · Mounting Screws
Page 41 of the Invacare XP6500 user manual — *HOLE# 1 2 3
Page 41 · click to enlarge
Page 41

SECTION 8—SEAT AND BACK

Adjusting the Back Height

1. Remove the two mounting screws and locknuts that secure the two back canes to the wheelchair frame.
2. Reposition the back canes to the desired height adjustment position.
3. Reinstall the mounting screws and locknuts that secure the back canes to the wheelchair frame and securely tighten.

Back Height (in inches)

*HOLE# 1 2 3

HEIGHT 17 18 19

*NOTE: Holes numbered from bottom to top for reference only. (There are no numbers on the back canes or wheelchair frame.)

FIGURE 8.3 Adjusting the Back Height

Show 5 more labels / short text on this page
Back Cane · Wheelchair Frame · Mounting · Locknut · Screw
Page 42 of the Invacare XP6500 user manual — SECTION 8—SEAT AND BACK
Page 42 · click to enlarge
Page 42

SECTION 8—SEAT AND BACK

Replacing Seat Positioning Strap

1. If necessary, remove the seat cushion from the seat upholstery.
2. Remove the two rear phillips screws and washers that secure the seat upholstery and seat positioning strap to the crossbraces.
3. Remove the two halves of the seat positioning strap from the crossbraces.
4. Position the two new seat positioning strap halves between the seat upholstery and the crossbraces.
5. Reinstall the rear phillips screws and washers that secure the seat upholstery and seat positioning strap to the crossbraces.
6. If necessary, reinstall the seat cushion onto the seat upholstery.

FIGURE 8.4 Replacing Seat Positioning Strap

Show 6 more labels / short text on this page
Washer · Rear Phillips Screw · Seat Upholstery · Seat Positioning · Strap · Crossbrace
Page 43 of the Invacare XP6500 user manual — Preparing the Joystick for Use
Page 43 · click to enlarge
Page 43

SECTION 9—JOYSTICK

SECTION 9—JOYSTICK

After ANY adjustments, repair or service and before use, make sure that all attaching hardware is tightened securely - otherwise injury or damage may result.

Before performing any maintenance, adjustment or service verify that On/Off switch on the joystick is in the Off position.

Preparing the Joystick for Use

1. Turn the adjustment lock lever to release the joystick mounting tube from the mounting bracket.
2. Slide joystick mounting tube to the desired position.
3. Turn the adjustment lock lever to secure the joystick mounting tube in the mounting bracket.

FIGURE 9.1 Preparing the Joystick for Use

Mounting Bracket · Joystick Mounting Tube · Adjustment Lock Lever

Page 44 of the Invacare XP6500 user manual — SECTION 9—JOYSTICK
Page 44 · click to enlarge
Page 44

SECTION 9—JOYSTICK

Repositioning the Joystick

1. Turn the adjustment lock lever to release the joystick mounting tube from the mounting bracket.
2. Remove the joystick from the wheelchair.
3. Remove the three hex screws that secure both halves of the mounting bracket to the arm tube.
4. Reposition mounting bracket on opposite arm tube ensuring the threaded plate of the mounting bracket is on the inside of the arm tube.
5. Using the three hex mounting screws and washers, secure both halves of the mounting bracket to the arm tube.
6. Slide the joystick mounting tube through the mounting bracket to the desired position.
7. Turn the adjustment lock lever to secure the joystick mounting tube into the mounting bracket.
8. If the tie-wraps securing the joystick control cord to the wheelchair were snipped at the beginning of this procedure, re-secure the cord to the wheelchair with new tie-wraps in such a way that the control cord will not be pinched or cut during operation of the wheelchair.

FIGURE 9.2 Repositioning the Joystick

Show 13 more labels / short text on this page
Mounting Bracket · (Threaded Plate) · Washers · Arm Tube · Hex Mounting · Screws · Adjustment · Mounting · Bracket · Lock Lever · Joystick · Joystick Mounting · Tube
Page 45 of the Invacare XP6500 user manual — Adjusting Forks
Page 45 · click to enlarge
Page 45

SECTION 10—CASTERS

SECTION 10—CASTERS

After ANY adjustments, repair or service and before use, make sure that all attaching hardware is tightened securely - otherwise injury or damage may result.

Before performing any maintenance, adjustment or service verify that On/Off switch on the joystick is in the Off position.

Adjusting Forks

1. Remove the dust cover from the caster headtube.
2. To properly tighten caster journal system and guard against flutter, perform the following check:

A. Remove battery boxes and battery tray from wheelchair. Refer to

Removing/Installing the Battery Boxes on page 49 and Installing/Removing Battery Tray on page 50.

B. Tip back the wheelchair to floor.

C. Pivot both forks and casters to top of their arc simultaneously.

D. Let casters drop to bottom of arc

(wheels should swing once to one-side, then immediately rest in a straight downward position).

E. Adjust locknuts according to

freedom of caster swing.

3. Test wheelchair for maneuverability.
4. Readjust locknuts if necessary, and repeat STEPS 1-3 until correct.
5. Snap dust cover into the caster headtube.

FIGURE 10.1 Adjusting Forks

Show 6 more labels / short text on this page
Dust Cover · Locknut · Nylon Washer · Caster Headtube · Fork · Caster
Page 46 of the Invacare XP6500 user manual — When to Charge Batteries
Page 46 · click to enlarge
Page 46

SECTION 11—BATTERIES

SECTION 11—BATTERIES

After ANY adjustments, repair or service and before use, make sure that all attaching hardware is tightened securely - otherwise injury or damage may result.

The use of rubber gloves is recommended when working with batteries. Before performing any maintenance, adjustment or service verify that On/Off switch on the joystick is in the Off position.

When to Charge Batteries

Recharge the batteries frequently. Do not wait until the battery charge is low to recharge. A battery’s life is extended if the charge level is maintained well above a low condition.

  • The information gauge is the primary source of user feedback. The true state-of-battery-charge notification is as follows:
  • Any GREEN LED lit: Batteries are at full or well charged.
  • Only AMBER and RED LEDs lit: Batteries are moderately charged. Recharge before undertaking a long trip.
  • Only RED LED lit: Battery charge is low. Recharge as soon as possible. NOTE: Accurate readings are displayed when the joystick is in neutral.

FIGURE 11.1 When to Charge Batteries

Information Gauge

Page 47 of the Invacare XP6500 user manual — TOOLS QUANTITY COMMENTS
Page 47 · click to enlarge
Page 47

SECTION 11—BATTERIES

Charging Batteries

Never attempt to recharge the batteries by attaching cables directly to the battery terminals or clamps. Use the recharging plug located on the front of the joystick. DO NOT attempt to recharge the batteries and operate the power wheelchair at the same time. During use and charging, unsealed batteries will vent hydrogen gas which is explosive in the right concentration with air. Follow the manufacturer's charging instructions.

The range per battery charge using recommended batteries should be approximately 5 to 9 hours of typical operation. Extensive use on inclines may substantially reduce per charge mileage.

Description and Use of Battery Chargers

The charger automatically reduces the charge from an initially high rate to a zero reading at a fully charged condition. If left unattended, the charger should automatically shut-off when full charge is obtained.

Some basic concepts which will help you understand this automatic process are:

The amount of electrical current drawn within a given time to charge a battery is called “charge rate”. If, due to usage, the charge stored in the battery is low, the charge rate is high, as indicated by the green light on the charger. Initially, the green light will stay illuminated for a short period of time followed by a longer period of off time. As a charge builds up, the charge rate is reduced, and the green light will stay illuminated for a longer period of time followed by a shorter off time. NOTE: If performing the charging procedures independently, READ and CAREFULLY follow the individual instructions for each charger (supplied or purchased).

TOOLS QUANTITY COMMENTS

Battery charger *Extension cord 1 1 Supplied Not Supplied

*NOTE3-prong plug, 15 ampere current rating; industrial type
Page 48 of the Invacare XP6500 user manual — CHAIR SEAT
Page 48 · click to enlarge
Page 48

SECTION 11—BATTERIES

Connecting the Battery Charger

1. Attach the battery charger connector to the charger port on the joystick.
2. Plug the charger’s AC power cord, or extension, into the grounded 120 VAC wall outlet.
3. Wait until charging is complete.

Replacing the Batteries

Most batteries are not sold with instructions. However, warnings are frequently noted on the cell caps. Read them carefully, otherwise injury or damage can occur. The use of rubber gloves is recommended when working with batteries.

The warranty and performance specifications contained in this manual are based on the use of deep cycle gel cell. ProBasics strongly recommends their use as the power source for this unit.

CHAIR SEAT

QTY VOLTS BATTERY

REMARKS

WIDTH

SIZE

18- inches 18-inches 2 2 12 12 U1 Group 22NF

Deep Cycle Deep Cycle

1. Remove the battery boxes and tray from the wheelchair. Refer to Removing/Installing the Battery Boxes on page 49 and Installing/Removing Battery Tray on page 50.
Show 4 more labels / short text on this page
Charger · Port · Charger · Recommended Battery Types
Page 49 of the Invacare XP6500 user manual — Removing/Installing the Battery Boxes
Page 49 · click to enlarge
Page 49

SECTION 11—BATTERIES

2. Remove the existing batteries from the battery boxes. Refer to Installing/Removing Batteries Into/From Battery Boxes on page 51.
3. Clean the new battery terminals.
4. Install the new batteries into the battery boxes. Refer to Installing/Removing Batteries Into/From Battery Boxes on page 51.
5. Install the battery boxes and tray onto the wheelchair. Refer to Removing/Installing the Battery Boxes on page 49 and Installing/Removing Battery Tray on page 50.

DO NOT allow the liquid in the battery to come in contact with skin, clothes or other possessions. It is a form of acid and harmful or damaging burns may result. Should the liquid touch your skin, wash the area immediately and thoroughly with cool water. In serious cases or if eye contact is made, seek medical attention IMMEDIATELY. NEVER smoke or strike a match near batteries. If the caps of the battery cells are removed, NEVER look directly into them while charging batteries.

1. Examine battery clamps and terminals for corrosion.
2. Verify that the plastic caps are in place over the battery cell holes.
3. Clean the terminals and inside the battery clamps by using a battery cleaning tool, wire brush, or medium grade sand paper.
4. Carefully dust off all metal particles.

Removing/Installing the Battery Boxes

WARNING NEVER unplug the battery cables by pulling on the wires. Unplug the battery cables by pulling on the connector ONLY. Otherwise injury or damage may result.

1. Verify that the On/Off switch on the joystick is in the Off position.
2. Place the front battery box onto the battery tray and connect controller connector to the front battery box. Refer to Electronics Manual (P/N 1123804).
3. Slide front battery box toward the front of the wheelchair.

Cleaning Battery Terminals

Page 50 of the Invacare XP6500 user manual — SECTION 11—BATTERIES
Page 50 · click to enlarge
Page 50

SECTION 11—BATTERIES

4. Place rear battery box onto the battery tray and connect the rear battery box to the front battery box.
Invacare XP6500 — SECTION 11—BATTERIES diagram (page 50)
Fig. 9
5. Connect the front battery box to the controller.

WARNING The battery box retaining strap MUST be fastened securely and adjusted prop- erly to hold battery boxes in place before using the wheelchair.

6. Connect the battery box retaining strap clip together and adjust if necessary.

FIGURE 11.3 Removing/Installing the

1. Attach the key slot bracket located at the front of the battery tray over the crossbrace bolt bushing under the seat upholstery of the wheelchair.
2. Attach the hanger brackets to the wheelchair frame.
Invacare XP6500 — SECTION 11—BATTERIES diagram (page 50)
Fig. 10

FIGURE 11.4 Installing/Removing Battery Tray

Show 13 more labels / short text on this page
To Controller · Front Battery Box · Battery · Tray · Rear Battery · Box Connector · Rear Battery · Retaining · Strap · Box · Battery Boxes · Installing/Removing Battery Tray · Hanger Bracket
Page 51 of the Invacare XP6500 user manual — TOOL QTY COMMENTS
Page 51 · click to enlarge
Page 51

SECTION 11—BATTERIES

Installing/Removing Batteries Into/From Battery Boxes

Make sure power to the wheelchair is Off before performing this procedure. The use of rubber gloves is recommended when working with batteries. ProBasics strongly recommends that battery installation and battery replacement ALWAYS be done by a qualified technician. After ANY adjustments, repair or service and before use, make sure all attaching hardware is tightened securely - otherwise injury or damage may result.

TOOL QTY COMMENTS

1/2-inch (6 pt.) Box Wrench 1 Not Supplied

7/16-inch (6pt) Box Wrench 1 Not Supplied

3/8-inch (6pt) Box Wrench 1 Not Supplied

Diagonal Cutters 1 Not Supplied

Always use the battery handle when lifting the battery. It is the most convenient method and assures that the battery acid will not spill. It also helps to prolong the life of the battery. DO NOT tip the batteries. Keep the batteries in an upright position.

1. If necessary, remove the battery boxes from the wheelchair. Refer to Removing/Installing the Battery Boxes on page 49.
2. Secure the battery lifting strap to battery terminal(s)/post(s).
Page 52 of the Invacare XP6500 user manual — DO NOT USE
Page 52 · click to enlarge
Page 52

SECTION 11—BATTERIES

3. Place batteries into battery box bottom.
Invacare XP6500 — SECTION 11—BATTERIES diagram (page 52)
Fig. 11

Molded Carrying Strap or Hold Down Flanges

U1 Battery Box Bottom

FIGURE 11.5 Installing/Removing Batteries Into/From Battery Boxes

Connecting Battery Cables

The use of rubber gloves is recommended when working with batteries. NEVER allow any of your tools and/or battery cable(s) to contact BOTH battery terminal(s)/post(s) at the same time. An electrical short may occur and serious personal injury or damage may occur. The correct battery terminal configurations shown in FIGURE 11.6 MUST be used. Batteries that have the reversed terminal configuration MUST NOT be used. Oth- erwise, serious injury or damage may occur.

DO NOT USE

NEGATIVE (-)

Invacare XP6500 — DO NOT USE diagram (page 52)
Fig. 12
Invacare XP6500 — DO NOT USE diagram (page 52)
Fig. 13

POSITIVE (+) Terminal/Post NEGATIVE (-)

FIGURE 11.6 Connecting Battery Cables

Show 4 more labels / short text on this page
POSITIVE (+) Terminal/Post · Terminal/Post · Terminal/Post · U1 Batteries
Page 53 of the Invacare XP6500 user manual — SECTION 11—BATTERIES
Page 53 · click to enlarge
Page 53

SECTION 11—BATTERIES

1. Install battery terminal cap(s) onto battery cable(s):

A. ORANGE battery terminal cap onto RED battery cable.

B. GREY battery terminal cap onto BLACK battery cable.

2. Connect battery cable(s) to battery(ies) terminal(s)/post(s) as shown in Detail “A”:

A. NEGATIVE (-) BLACK battery cable to NEGATIVE (-) battery terminal/post.

B. POSITIVE (+) RED battery cable to POSITIVE (+) battery terminal/post.

3. Secure the battery cable(s)/ring terminal(s) to the battery terminal(s)/post(s), BLACK to NEGATIVE (-) and RED to POSITIVE (+), with the provided 1/4-20 x 7/8-inch hex flange screw and hex flange locknut as shown in Detail “A”. Securely tighten.
4. Verify all battery cable(s)/ring terminal(s) are correctly installed and securely tightened.
5. Slide terminal cap(s) down battery cable(s) and onto battery clamps.
6. Using the 11½-inch long tie-wrap, secure each terminal cap in place (Detail “A”).
7. Install the battery box top(s).
8. Install the battery box(es) into the wheelchair. Refer to Removing/Installing the Battery Boxes on page 49.
9. If necessary, charge the battery(ies). Refer to Charging Batteries on page 47.
Page 54 of the Invacare XP6500 user manual — NEGATIVE (-)
Page 54 · click to enlarge
Page 54

SECTION 11—BATTERIES

Detail “A” - Dual UI Batteries

1/4-20 x 7/8-Inch Hex Flange Screw

NEGATIVE (-)

NEGATIVE (-)

Invacare XP6500 — NEGATIVE (-) diagram (page 54)
Fig. 14

1/4-20 Hex Flange Locknut

Ring Terminal to This Side of Terminal/Post

ORANGE Terminal Cap Tie-Wraps

Detail “B” - Installing Battery Terminal Cap onto Battery Cable

FIGURE 11.7 Connecting Battery Cables

Replacing Battery Box Retaining Strap

WARNING The battery box retaining strap MUST be fastened securely and adjusted properly to hold battery boxes in place before using the wheelchair.

1. Remove the battery boxes from the wheelchair. Refer to Removing/Installing the Battery Boxes on page 49.
2. Remove the battery tray from the wheelchair. Refer to Installing/Removing Battery Tray on page 50.

WARNING The retaining strap MUST be fed through the adjustable end of the battery box retaining strap buckle as shown in FIGURE 11.8.

Show 7 more labels / short text on this page
Terminal/Post · POSITIVE (+) Terminal/Post · Terminal/Post GREY Terminal · POSITIVE (+) Terminal/Post · Cap · Connect Cable · Cable Terminal Cap
Page 55 of the Invacare XP6500 user manual — SECTION 11—BATTERIES
Page 55 · click to enlarge
Page 55

SECTION 11—BATTERIES

3. Feed the existing battery box retaining strap through the adjustable end of the battery box retaining strap buckle.
4. Feed the existing battery box retaining strap through the slots in the battery box tray.
5. Install the new battery box retaining strap by reversing STEPS 3 and 4.
6. Reinstall the battery tray from the wheelchair. Refer to Installing/Removing Battery Tray on page 50.
7. Reinstall the battery boxes from the wheelchair. Refer to Removing/Installing the Battery Boxes on page 49.

FIGURE 11.8 Replacing Battery Box Retaining Strap

Show 14 more labels / short text on this page
Adjustable End · Retaining · Strap Buckle · Retaining · Battery · Strap · Tray · Slots · Retaining · Slots · Strap · Battery Tray · Retaining · Strap
Page 56 of the Invacare XP6500 user manual — Engaging/Disengaging the Clutches
Page 56 · click to enlarge
Page 56

SECTION 12—CLUTCH/MOTOR LOCK

SECTION 12—CLUTCH/MOTOR LOCK

WARNING After ANY adjustments, repair or service and before use, make sure all attaching hardware is tightened securely - otherwise injury or damage may result.

Engaging/Disengaging the Clutches

WARNING DO NOT engage or disengage the clutches until the power is Off.

The clutch engagement/disengagement allows freewheeling or joystick controlled operation. Freewheeling allows an attendant to maneuver the wheelchair without power. To engage/disengage the clutches:

1. Locate the clutch handles on motors.

WARNING If clutch handles are forced to engage in the wrong direction as shown in FIGURE 12.1, the motors will be damaged and will need to be replaced.

2. Perform one of the following:
  • To Engage: turn the clutch handles until they are pointing towards the rear of the wheelchair. NEVER try to turn the clutch handles towards the front of the wheelchair.
  • To Disengage: turn the clutch handles until they are pointing towards the rear wheels. NEVER try to turn the clutch handles towards the inside of the wheelchair.

Top View Of Wheelchair

Invacare XP6500 — Engaging/Disengaging the Clutches diagram (page 56)
Fig. 15 · Engaging/Disengaging the Clutches

FIGURE 12.1 Engaging/Disengaging the Clutches

Page 57 of the Invacare XP6500 user manual — Installing/Adjusting/Using the Wheel Locks
Page 57 · click to enlarge
Page 57

SECTION 13—WHEEL LOCKS

SECTION 13—WHEEL LOCKS

WARNING After ANY adjustments, repair or service and before use, make sure all attaching hardware is tightened securely - otherwise injury or damage may result.

Installing/Adjusting/Using the Wheel Locks

1. Position the wheel lock on the wheelchair frame.
2. Loosely install the hex screw and locknut that secures the wheel lock to the wheelchair frame.
3. Make sure wheel lock is disengaged from rear wheel.
4. Measure the distance between the wheel lock shoe and the rear wheel.
5. Slide the wheel lock along the wheelchair until the measurement is between 5/32 and 5/16-inches.
6. Tighten the wheel lock to the wheelchair frame.
7. Repeat STEPS 2-6 for the opposite wheel lock.
8. Disengage the clutches. Refer to Engaging/Disengaging the Clutches on page 56.
9. Engage the wheel locks and push against the wheelchair to determine if the wheel locks engage the rear wheels enough to hold the wheelchair.
10. Repeat STEPS 3-9 until the wheel locks engage the rear wheels enough to hold the wheelchair.
11. Engage clutches. Refer to Engaging/Disengaging the Clutches on page 56.

Installing/Adjusting

Page 58 of the Invacare XP6500 user manual — SECTION 12—CLUTCH/MOTOR LOCK
Page 58 · click to enlarge
Page 58
Invacare XP6500 — SECTION 12—CLUTCH/MOTOR LOCK diagram (page 58)
Fig. 16

The wheelchair is equipped with a pair of independently operated wheel locks located just in front of the rear wheels.

1. To engage the wheel locks, grip the handle and push forward to the lock position.
Page 59 of the Invacare XP6500 user manual — Installing/Removing/Adjusting Anti-Tippers
Page 59 · click to enlarge
Page 59

SECTION 14—ANTI-TIPPERS

SECTION 14—ANTI-TIPPERS

WARNING After ANY adjustments, repair or service and before use, make sure all attaching hardware is tightened securely - otherwise injury or damage may result.

Installing/Removing/Adjusting Anti-Tippers

Anti-tippers MUST be fully engaged and spring buttons fully protruding out of adjustment holes before using the wheelchair. Ensure both anti-tippers have the same ground clearance. A 1½ to 2-inch clearance between the bottom of the anti-tipper wheels and the ground/floor MUST be maintained at all times. Anti-tippers MUST be used at all times. When outdoors on wet, soft ground or gravel surfaces, anti-tippers may not provide the same level of protection against tipover. Extra caution must be observed when traversing such surfaces.

1. Press the two release buttons in and insert the anti-tippers with anti-tipper wheels pointing toward the ground/floor into the support tubes until the release buttons lock in place.
2. Measure the distance between the bottom of the anti-tipper wheels and the ground/floor and ensure that it is 1½-2-inches. If necessary, adjust height. To adjust the height of the anti-tippers, refer to Adjusting Height on page 59.
1. Press release buttons in and remove the anti-tippers from the support tubes.
1. Press the release buttons in and adjust the anti-tippers to the desired height.

Installing · Removing · Adjusting Height

Page 60 of the Invacare XP6500 user manual — SECTION 14—ANTI-TIPPERS
Page 60 · click to enlarge
Page 60

SECTION 14—ANTI-TIPPERS

Invacare XP6500 — SECTION 14—ANTI-TIPPERS diagram (page 60)
Fig. 17

FIGURE 14.1 Installing/Removing/Adjusting Anti-Tippers

Page 61 of the Invacare XP6500 user manual — Transporting the Wheelchair
Page 61 · click to enlarge
Page 61

SECTION 15—TRANSPORTING

SECTION 15—TRANSPORTING

WARNING After ANY adjustments, repair or service and before use, make sure all attaching hardware is tightened securely - otherwise injury or damage may result.

Transporting the Wheelchair

WARNING The weight of the wheelchair without the user and batteries is approximately 82 lbs. Use proper lifting techniques (lift with your legs) to avoid injury. Extreme caution is advised when it is necessary to move an unoccupied power wheelchair. ProBasics recommends using two assistants and making thorough preparations. Make sure to use ONLY secure, non-detachable parts for hand-hold supports.

Keep hands and fingers clear of moving parts to avoid injury. DO NOT place hand or fingers on the underside of the seat rail when opening or closing the wheelchair. DO NOT sit or transfer into the wheelchair unless it is fully open and the seat rails are fully seated into the side frame H-blocks.

1. Push down on the seat rails until the seat rails are positioned in the seat guides.
2. Assemble the wheelchair by following the instructions in this manual.
1. Remove the anti-tippers. Refer to Installing/Removing/Adjusting Anti-Tippers on page 59.
1. Remove the batteries. Refer to Removing/Installing the Battery Boxes on page 49.
2. Remove the footrests/legrests from the wheelchair. Refer to Installing/Removing/Using the Footrest/Legrest on page 33.
3. Remove the battery tray from the wheelchair. Refer to Installing/Removing Battery Tray on page 50.
4. Lift up on the seat rails of the wheelchair.

Unfolding/Folding the Wheelchair · Unfolding · Folding

Page 62 of the Invacare XP6500 user manual — TRANSPORTING
Page 62 · click to enlarge
Page 62

TRANSPORTING

Invacare XP6500 — TRANSPORTING diagram (page 62)
Fig. 18

Refer to Removing/Installing the Battery Boxes on page 49

FIGURE 15.1 Transporting the Wheelchair

Page 63 of the Invacare XP6500 user manual — LIMITED WARRANTY
Page 63 · click to enlarge
Page 63

LIMITED WARRANTY

LIMITED WARRANTY

PLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975.

This warranty is extended only to the original purchaser/user of our products.

This warranty gives you specific legal rights and you may also have other legal rights which vary from state to state.

With regards to the original purchaser/user only, ProBasics warrants the front and rear frames to be free from defects in materials and workmanship for a period of three years from date of purchase; seat frame for a period of three years from the date of purchase; electronics, motors and gearboxes for a period of one year from the date of purchase; all remaining components for one year from the date of purchase except upholstered materials, padded materials and tires/wheels. If within such warranty period any product shall be proven to be defective, such product shall be repaired or replaced, at ProBasics' option. This warranty does not include any labor or shipping charges incurred in replacement part installation or repair of any such product. ProBasics' sole obligation and your exclusive remedy under this warranty shall be limited to such repair and/or replacement.

For warranty service, please contact the dealer from whom you purchased your ProBasics product. In the event you do not receive satisfactory warranty service, please contact ProBasics at the address on the bottom of the back cover. Provide dealer's name address, date of purchase, indicate nature of the defect and, if the product is serialized, indicate the serial number. Do not return products to ProBasics without our prior consent.

Limitations and Exclusions: The foregoing warranty shall not apply to serial numbered products if the serial number has been removed or defaced, products subject to negligence, accident, improper operation, maintenance or storage, commercial or institutional use, products modified without ProBasics' express written consent including, but not limited to, modification through the use of unauthorized parts or attachments; products damaged by reason of repairs made to any component without the specific consent of ProBasics, or to a product damaged by circumstances beyond ProBasics' control, and such evaluation will be solely determined by ProBasics. The warranty shall not apply to problems arising from normal wear or failure to adhere to these instructions. A change in operating noise, particularly relative to motors and gearboxes does not constitute a failure. All such devices will exhibit a change in operating noise due to aging.

The foregoing express warranty is exclusive and in lieu of any other warranties whatsoever, whether express or implied,

including the implied warranties of merchantability and fitness for a particular purpose, and the sole remedy for violations of any warranty whatsoever, shall be limited to repair or replacement of the defective product pursuant to the terms contained herein, the application of any implied warranty whatsoever shall not extend beyond the duration of the express warranty provided herein. ProBasics shall not be liable for any consequential or incidental damages whatsoever.

THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE/PROVINCIAL LAWS AND REQUIREMENTS.

Page 64 of the Invacare XP6500 user manual — SECTION 12—CLUTCH/MOTOR LOCK
Page 64 · click to enlarge
Page 64

ProBasics is a trademark of PMI, Inc.

Invacare XP6500 — SECTION 12—CLUTCH/MOTOR LOCK diagram (page 64)
Fig. 19

!"#$!#% !"#$!#%

Distributed by ProBasics. Part No. XP6500-OM/1123839 Rev C - 08/18/06

Fix-it

Troubleshooting

Common problems, likely causes, and what to do.

ProblemPossible causeRemedy
Batteries draw excessive current when charging.Battery failure. Electrical malfunction.Have batteries checked for shorted cell. Replace if necessary. Contact dealer for service. p.26
Battery indicator flashes the charge level is low immedi- ately after recharging.Battery failure. Charger malfunction. Electrical malfunction.Check batteries for shorted cell. Replace if nec- essary. Contact dealer/ProBasics. Poor connections between charger and wheel- chair. Contact dealer for service. p.26
Battery indicator flashes the charge level is low too soon after being recharged.Batteries not charged. Weak batteries. Electrical malfunction.Have charger checked. Replace batteries if necessary. Contact dealer for service. p.26
Motor chatters or runs irregularly.Electrical malfunction.Contact dealer. p.26
Wheelchair does not respond to commands. Power on, battery indicator flashes.One or both clutches disen- gaged.Engage motor locks/clutches. p.27
Only one drive wheel turns.Electrical malfunction. One clutch is disengaged.Contact dealer for service. Engage clutch. p.27
Joystick erratic or does not respond as desired.Electrical malfunction. Controller programmed improperly.Contact dealer for service. Reprogram controller. Contact dealer for ser- vice. p.27
Wheelchair does not respond to commands. Power indicator off - even after recharging.Poor battery terminal con- nection. Electrical malfunction.Clean terminals. Contact dealer. p.27

Ask about the Invacare XP6500

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Invacare XP6500

Be the first to ask — other owners or our team will answer.

Original document

Invacare XP6500 manual PDF

64 pages·2.1 MB·Mobility Aid·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…