IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: 61-340 MODEL: 61-342 Multimeter Service Information The Service Information provides the following information: • Precautions and safety information .
Form Number 2 TM61340-2 Rev November 2007 TABLE OF CONTENTS Page Introduction 1 Precautions and Safety Information 1 Symbols 1 Safety 2 S pecifications 3 General S pecifications 3 Measurement Characte.
Page 1 Introduction W arning T o avoid shock or injury , do not perform the verification test s or calibration procedu res described in the manual unless you are qualified to do so. The information provided in this document is for the use of q ualified personnel only .
Page 2 SAFETY Review the following safety precautions to avoid injury and prevent damag e to this product or any product s connected to it. T o avoid potential hazards, use the product only as specified. For operating instructions, see the 61 -340 / 61-342 Digit al Multimeter Instruction Manual.
Page 3 SPECIFICATIONS All specifications are warranted unles s noted typical and apply to the 61-340 & 61-342 S tated accu racies are at 23°C±5°C at less than 75% relative humidity and without the battery indicator displayed.
Form Number 2 TM61340-2 Rev November 2007 Page 4 Measurement Characteristics Accuracy is ±(% reading + number of digits ) at 23°C ± 5°C, less tha n 75% R.H. (1) DC Volts(for 61-340 / 61-342) Range Resolution Accuracy Over volt age protection 400.0mV 0.
Form Number 2 TM61340-2 Rev November 2007 Page 5 (3b) DC Current (61-340 / 61-342) Range Resolution Accuracy Input Protection 4.000A 0.001A 10.00A* 0.01A ±(1.5% reading + 5 digit s) 10A, 250V Fast Blow Fuse Overload Protection: A Input: 10A, 250V Fast Blow fuse.
Page 6 (5) Resist ance (for 61-340 / 61-342) Range Resolution Accuracy Over volt age protection 400.0 Ω * 1 0.1 Ω ±(1.2% reading + 5 digit s) 4.000K Ω 1 Ω 40.00K Ω 10 Ω 400.0K Ω 100 Ω ±(1.0% reading + 2 digit s) 4.000M Ω 1K Ω ±(1.
Form Number 2 TM61340-2 Rev November 2007 Page 7 (8) Frequency (for 61-340 / 61-34 2) Range Resolution Sensitivity Accuracy Overload protection 10.00Hz 0.01Hz 100.0Hz 0.1Hz 1.000KHz 1Hz 10.00KHz 10Hz 100.0KHz 100Hz 1.000MHz 10KHz 0.7V rms * 4.000MHz 10KHz 3.
Form Number 2 TM61340-2 Rev November 2007 Page 8 (14) REL Press this button the meter enter relative measuring st ate and “ Δ ” symbol appear on LCD. The result of the relative measurement is the dif ference between measuri ng value and reference value.
Form Number 2 TM61340-2 Rev November 2007 Page 9 Physical and Environment al Characteristics Characteristics Description Dimensions (H×W ×D) 180mm(H) ×91mm (W) ×43mm(D) (with holste r) 7.1” (H) x 3.6”(W) x1.7”(D) Weight (with battery& holster) 0.
Form Number 2 TM61340-2 Rev November 2007 Page 10 Required Equipment Required equipment is listed in T able B. If the recommended model s are not available, equipment with equivalent specifications may be used. Repairs or se rvicing should be per formed only by qualified personnel.
Page 1 1 Basic Maintenance W arning T o avoid shock, remove the test leads and any input signals before opening the case or re placing the battery or fuses. Opening the Meter Case Caution T o avoid unintentional shock circuit, alwa ys place the uncovered meter as sembly on a protective surface.
Page 12 Replacing Fuses W arning To avoid electrical shock, remove the test lead s and any input signals b efore replacing the battery or fuses. To prevent damage or inju ry, INSTA LL ONLY quick actin.
Page 13 Performance Tests The following performance test s verify the complete operability of the meter and check the accura cy of each meter function against the meter’s specifications. For operating instructions, see the 61 -340 / 61-342 Digit al Multimeter Instruction Manual.
Page 14 T esting the V olt age Function (for 61-340 / 61-342) To verify accuracy in the AC and DC voltage ran ges, do the following: 1. Turn the rotary switch to “ ” position. 2. Connect the calibrator to the and COM inputs on the meter. 3. Set the calibrator for the voltage from step 1 to 7 in Table 1.
Page 15 Table 2 AC Voltage Test: Step Range Input Frequency Reading 61-340 Reading 61-342 1 400mV 300mV 60 295.6 to304.4 295.6 to304.4 2 400mV 300mV 400 295.6 to304.4 295.6 to304.4 3 4V 3.000V 60Hz 2.965 to 3.035 2.965 to 3.035 4 4V 3.000V 400Hz 2.965 to 3.
Page 16 Table 3b DC mA Test: Step Range Source Reading 1 40mA 30.00mA 29.73 to 30.27 2 400mA 300.0mA 297.3 to 302.7 T esting the DC ampere Function (fo r 61-340 and 61-342) 1. Connect the calibrator to the 10A and COM inputs on the meter . 2. T urn the rotary switch to .
Page 17 T esting the AC milliampere function: (for 61-3 40and 61-342) 1. Connect the calibrator to the and COM inputs on the meter . 2. T urn the rotary switch to . a. Press the SELECT button to select the AC function b. Press the REL button to make the display is zero 3.
Page 18 T esting the Resist ance Function (for 61-340 / 61-342) T o verify the accuracy of the re sistance function, do the following: 1. Connect the calibrator to and COM on the meter . 2. T urn the rotary switch to . a. Press the SELECT button to select the Ω function 3.
Page 19 Table 6 Capacitan ce Test: Step Range Source Reading 1 40nF 10.000nF 9.60 to 10.40 2 40nF 30.00nF 29.00 to 31.00 3 400nF 300.0nF 290.5 to 309.5 4 4µF 3.
Page 20 T esting the Frequency Function (for 61-340 / 61-3 42) T o verify the accuracy of the me ter’s frequency function, do the following: 1. Connect the calibrator to the and COM inputs on the meter . 2. Note: The accuracy of the calibrato r’s frequency function must be appropriate for the specified accuracy of the meter.
Page 21 T esting the T emperature Function (fo r 61-340 / 61-342) T o verify the accuracy of the meter ’ s temperature function, do the following: 1. Connect the calibrator to the and COM input s on the meter . 2. Set the rotary switch to the º F / º C as instructed in Table 8.
Page 22 Calibration Procedure Recalibrate your meter: It is recommended that the multimeter be calibrated on ce each year . 1. Perform calibration at an a mbient temperature of 23°C±2°C and a relative humidity of <70% Disconnect the test leads and turn the m eter off.
Page 23 6. Using a small flat-tipped screwdrive r , adjust VRtr until the display reads 2.995 to 3.005V olts 7. Disconnect the AC calibrator from the meter . (C) TEMP Calibration (Adjust VR3 ) 1. Set the rotary switch to the " º F/ º C " position.
An important point after buying a device IDEAL INDUSTRIES 61-340 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought IDEAL INDUSTRIES 61-340 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data IDEAL INDUSTRIES 61-340 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, IDEAL INDUSTRIES 61-340 you will learn all the available features of the product, as well as information on its operation. The information that you get IDEAL INDUSTRIES 61-340 will certainly help you make a decision on the purchase.
If you already are a holder of IDEAL INDUSTRIES 61-340, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime IDEAL INDUSTRIES 61-340.
However, one of the most important roles played by the user manual is to help in solving problems with IDEAL INDUSTRIES 61-340. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device IDEAL INDUSTRIES 61-340 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center