__________ *This bulletin supersedes TB 9 - 6625 - 2049 - 35, 15 January 1996. *TB 9 - 6625 - 2049 - 35 SUPERSEDED COPY DATED 15 JANUARY 1996 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROC.
TB 9 - 6625 - 2049 - 35 2 Paragraph Page SECTION II I. CALIBRATION PROCESS - Continued Triangle and ramp amplitude ............................ 16 17 Dc offset ............................................................... 17 18 Dc of fset with ac ..
TB 9 - 6625 - 2049 - 35 3 Table 1 . Calibration Description Test instrument parameters Performance specifications Harmonic distortion 1 (relative to fundam ental at full output) Fundamental frequency: 0.
TB 9 - 6625 - 2049 - 35 4 Table 1. Calibration Description - Continued Test instrum ent parameters Performance specifications Amplitude accuracy with dc offset and no attenuation (range 1) into 50 Ω load Tolerance relative to programmed amplitude Sine wave: .
TB 9 - 6625 - 2049 - 35 5 5. Accessories Required. The accessories required for this calibration are common usage accessories, issued as indicated in paragraph 4 above, and are not listed in this calibration procedure.
TB 9 - 6625 - 2049 - 35 6 SECTION III CALIBRATION PROCESS 6. Preliminary Instructions a . The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration process. Personnel should become familiar with the entire bulletin b efore beginning the calibration.
TB 9 - 6625 - 2049 - 35 7 8. Harmonic Distortion a. Performance Check (1) Connect TI EXT REF IN 1, 10 MHz to spectrum analyzer 10 MHz REF IN/OUT . (2) Connect TI SIGNAL to spectrum analyzer INPUT 50 Ω Ω . (3) Press keys and enter values using DATA keys as listed in (a) through (h) below: (a) SIGNAL off (option 002 not lit).
TB 9 - 6625 - 2049 - 35 8 (a) 50 Hz . (b) ENTRY AMPTD . (c) 10 VOLT . (13) Set audio analyzer to measure distortion in dB. Audio analyzer indication will be ≤ - 65 dB. (14) Perform (15) through (19) below for option 002 only . (15) Connect equipm ent as shown in figure 1 and set resistance standard to 470 Ω .
TB 9 - 6625 - 2049 - 35 9 9. Amplitude Modulation Distortion a. Performance Check (1) Press keys and enter values using DATA keys as listed in (a) throug h (i) below: (a) SIGNAL off (option 002 not lit). (b) FUNCTION sine wave. (c) ENTRY FREQ . (d) 1 MHz .
TB 9 - 6625 - 2049 - 35 10 (a) SIGNAL off (option 002 not lit). (b) FUNCTION square wave. (c) ENTRY FREQ . (d) 1 MHz . (e) ENTRY AMPTD . (f) 1 V RMS . (2) Connect TI SIGNAL to oscilloscope CH 1 . (3) Set oscilloscope CH 1 COUPLING 50 Ω Ω to ON . (4) Set osc illoscope controls for duty cycle measurement.
TB 9 - 6625 - 2049 - 35 11 (4) Set oscilloscope controls to measure ramp retrace time from the 90 to 10 percent points. Ramp retrace time will be ≤ 3 µ s. ( 5) Press FUNCTION negat ive ramp key and repeat (4) above. b. Adjustments . No adjustments can be made.
TB 9 - 6625 - 2049 - 35 12 (15) Set frequency counter controls to measure frequency. If frequency counter does not indic ate between 59.99970 and 60.00030 MHz, perform b below. b. Adjustments (1) Disconnect TI rear panel 10 MHz OVEN OUTPUT from EXT REF IN option 001 only.
TB 9 - 6625 - 2049 - 35 13 (12) Adjust A9R7 Fine Adj. (fig. 3) and A9 Coarse Adj. (fig. 3) for a minimum indication on frequency difference meter. (13) Disconnect TI from frequency difference meter. (14) Reconnect TI rear panel 10 MHz OVEN OUTPUT to EXT REF IN .
TB 9 - 6625 - 2049 - 35 14 (6) Adj ust frequency counter sample rate control to HOLD and press RESET pushbutton. Record frequency counter indication. (7) Press ENTRY PHASE key and enter - 1 deg using DATA keys. (8) Press frequency counter RESET pushbutton and record frequency counter indi cation.
TB 9 - 6625 - 2049 - 35 15 Table 3. Amplitude Test instrument settings Multimeter indications ENTRY AMPTD ENTRY FREQ (V ac) 3.536 V RMS 1 kHz 3.495 3.577 3.536 V RMS 100 kHz 3.495 3.577 1.061 V RMS 100 kHz 1.048 1.073 1.061 V RMS 1 kHz 1.048 1.073 1.061 V RMS 100 Hz 1.
TB 9 - 6625 - 2049 - 35 16 Table 4. Flatness Test instrument and measuring receiver setting s (MHz) 1 2 4 6 8 10 12 14 16 18 20 b. Adjustments . No adjustments can be made. 15. Square Wave Amplitude and Flatness a. Performance Check (1) Connect TI SIGNAL to multimeter INPUT HI and LO using 50 Ω feedthrough termination.
TB 9 - 6625 - 2049 - 35 17 (a) 1 kHz . (b) ENTRY AMPTD . (c) 1 V RMS . (d) AMPTD CAL . (7) Multimeter will indicate between 0.99 and 1.01 V ac. (8) Disconnect TI from multimeter. (9) Connect TI SIGNAL to oscilloscope CH 1 using 50 Ω feedthrough termination.
TB 9 - 6625 - 2049 - 35 18 (6) Enter 10 kHz using DATA keys. Multimeter will indicate between 2.742 and 3.032 V ac. (7) Press keys and enter values using DATA keys as listed in (a) through (c) below: (a) FUNCTION positive ramp. (b) 100 Hz . (c) AMPTD CAL .
TB 9 - 6625 - 2049 - 35 19 Table 5. Dc Offset - Continued Test instrument DC OFFSET Multimeter indications (V dc) se ttings Min Max 14.99 mV 0.01493 0.01505 - 14.99 mV - 0.01505 - 0.01493 4.999 mV 0.004979 0.005019 - 4.999 mV - 0.005019 - 0.004979 1.499 mV 0.
TB 9 - 6625 - 2049 - 35 20 (4) Enter - 4.5 VOLT using DATA keys. Multimeter will indicate between - 4.35 and - 4.65 V dc. (5) Press ENTRY FREQ key and enter 999.9 kHz using DATA keys. Multimeter will indicate between - 4.44 and - 4.56 V dc. (6) Press ENT RY DC OFFSET key and enter 4.
TB 9 - 6625 - 2049 - 35 21 (a) ENTRY FREQ . (b) 100 kHz . (c) ENTRY AMPTD . (d) 10 V RMS . (e) AMPTD CAL . (5) Multimete r will indicate between 9 and 11 V ac. (6) Press keys and enter values using DATA keys as listed in (a) through (f) below: (a) FUNCTION square wave.
TB 9 - 6625 - 2049 - 35 22 Figure 5. Power supply assembly A2 – test instrument top view. (5) Press presently active FUNCTION key to remove ac output and activate ENTRY DC OFFSET key (lit). (6) Enter 0 VOLT using DATA keys and press AMPTD CAL key. (7) Connect multimeter INPUT HI to TP AMP OUT (fig.
TB 9-6625-2049-35 By Order of the Secretary of the Army: Official: ERIC K. SHINSEKI General, United States Army Chief of Staff JOEL B. HUDSON Administrative Assistant to the Sec retary of the Army 0029703 DISTRIBUTION: To be distributed in accordance with IDN 342200, requirements for calibration procedure TB 9-6625-2049-35.
An important point after buying a device HP (Hewlett-Packard) 3325B (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought HP (Hewlett-Packard) 3325B yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data HP (Hewlett-Packard) 3325B - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, HP (Hewlett-Packard) 3325B you will learn all the available features of the product, as well as information on its operation. The information that you get HP (Hewlett-Packard) 3325B will certainly help you make a decision on the purchase.
If you already are a holder of HP (Hewlett-Packard) 3325B, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime HP (Hewlett-Packard) 3325B.
However, one of the most important roles played by the user manual is to help in solving problems with HP (Hewlett-Packard) 3325B. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device HP (Hewlett-Packard) 3325B along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center