Instruction/ maintenance manual of the product JB0160-010-220 Graham Field
Go to page of 28
OXYGEN CONCENTRA TOR USER MANUAL Model Numbers JB0160-010B, JB0160-010, JB0160-010B-220, JB0160-010-220 and JB0160-015B, JB0160-015, JB0160-015B-220, JB0160-015-220 with O 2 Sensor Impor tant: Do not operate this concentrator without rst reading and understanding this manual! Save this manual for future use.
2 John Bunn O2 Premier Oxygen Concentrator All specications and pr oduct congurations ar e subject to change without notication. CONTENTS 1 SAFETY ................................................................................... 3 Important Information .
3 John Bunn O2 Premier Oxygen Concentrator 1 SAFETY The John Bunn O2 Pr emier oxygen concentrator is intend- ed for use as an oxygen supplement device in a home or car e facility . Use the concentrator only as prescribed by a physician. This user manual ser ves as a refer ence for your oxygen concentrator .
4 John Bunn O2 Premier Oxygen Concentrator s Damage: If the electrical power source becomes unstable, discontinue use and find an alternate source. s Damage: The oxygen concentrator cabinet contains no ser viceable par ts! The oxygen concentrator cabinet should ONL Y be opened by an authorized equipment provider .
5 John Bunn O2 Premier Oxygen Concentrator W ARNING: T o reduce the risk of fire, special care must be taken when using oxygen concentrators. DO NOT SMOKE while using this device. K eep all matches, lighted cigarettes or other sources of ignition out of the room in which this concentrator is located.
6 John Bunn O2 Premier Oxygen Concentrator Additional Safety W ar nings W ARNING: TO REDUCE THE RISK OF BURNS, ELECTROCUTION, FIRE OR INJURY TO PERSONS, avoid using while bathing. If continuous usage is required by the direction of a ph y sician, the concentrator must be located in another room at least 9 feet (2.
7 John Bunn O2 Premier Oxygen Concentrator A TTENTION: AL W A YS super vise closely when this product is used near children or those who require close super vision. A TTENTION: Only use this device as intended and described in this manual. A TTENTION: DO NOT use par ts, accessories or adapter s other than those authorized by manufacturer .
8 John Bunn O2 Premier Oxygen Concentrator 2 FEA TURES Summary The John Bunn O 2 Pr emier oxygen concentrator is intend- ed for use as an oxygen supplement device in a home or car e facility .
9 John Bunn O2 Premier Oxygen Concentrator 3 HANDLING GF Health Products, Inc. Fr eight Policy For Y our Pr otection, Read Carefully The carrier accepted this mer chandise “in good condition” and is r esponsible for safe deliver y . Befor e signing the fr eight bill, inspect the shipment car e- fully for damage or missing pieces.
10 John Bunn O2 Premier Oxygen Concentrator Inspection Examine exterior of the oxygen concentrator for nicks, dents, scratches or other damages. Inspect all compo- nents. Storage Stor e the r epackaged oxygen concentrator in a dr y area. DO NOT place anything on top of the r epackaged concen- trator .
11 John Bunn O2 Premier Oxygen Concentrator 4 OPERA TION AND INST ALLA TION Preparing the Concentrator (refer to above pictur e) 1. Unscr ew the humidier bottle cover , if so equipped (hu- midier bottle is optional). 2. Fill the humidier bottle with puried water (or dis- tilled water) between the maximum and minimum water level lines.
12 John Bunn O2 Premier Oxygen Concentrator trator’ s AC connector into a properly gr ounded power outlet. T ur ning the Concentrator On 1. Pr ess the power switch to start the concentrator . There will be a brief delay befor e the concentrator starts.
13 John Bunn O2 Premier Oxygen Concentrator 5 MAINTENANCE Cleaning the Cabinet Clean the cabinet at least once a month. 1. T urn of f the power switch. 2. Clean the cabinet with a mild household cleaner and a non-abrasive cloth or sponge. Do not spill liquid inside the cabinet.
14 John Bunn O2 Premier Oxygen Concentrator Side filters Note: For earlier models only , you will need a small blade screwdriver to open the side filter access door . For later models, the filter grill simply pulls out. If this operation is too dif ficult for you to per form, please contact your equipment provider to perfor m this for you.
15 John Bunn O2 Premier Oxygen Concentrator • A wood stove is used • Ker osene or oil is used • Pets ar e present • T obacco is used Intake filter Note: Please have your Graham-Field equipment provider replace the intake filter . This operation requires opening the intake filter access door .
16 John Bunn O2 Premier Oxygen Concentrator 6 TROUBLESHOOTING GUIDE SYMPTOM CHECK / ACTION Concentrator power is Check if concentrator is unplugged; plug in. on and alarm is on, but concentrator doesn’t Check if circuit br eaker is tripped; x circuit work breaker .
17 John Bunn O2 Premier Oxygen Concentrator SYMPTOM CHECK / ACTION Concentrator doesn’t Discontinue use; contact Graham-Field equipment work; alarm sounds; provider immediately .
18 John Bunn O2 Premier Oxygen Concentrator 7 BILL OF MA TERIALS Y our concentrator includes the following components: • Oxygen concentrator (one each) • User manual (one each) • Humidier bot.
19 John Bunn O2 Premier Oxygen Concentrator 8 SPECIFICA TIONS • Flow Range: 1 to 5L/min • Oxygen Concentration: 1-4 LPM 94% ± 2% 5 LPM 92% ± 2% • Maximum Outlet Pressur e: 6.52 psi ± .065 psi (45kPa ± 4.5kPa) • Pressur e Relief Mechanism Operational at: 36.
20 John Bunn O2 Premier Oxygen Concentrator 9 LIMITED W ARRANTY This warranty is extended only to the original pur chaser of our pr oducts. This warranty gives you specic legal rights and you may also have other legal rights which may var y fr om state to state.
21 John Bunn O2 Premier Oxygen Concentrator as identication, within thirty (30) days of the return autho- rization date. Do not r eturn pr oducts to GF Health Pr oducts, Inc. or its Authorized Equipment Pr oviders, without prior consent. C.O.D. shipments will be r efused.
22 John Bunn O2 Premier Oxygen Concentrator 10 INDEX A A TTENTION, signicance 3 B Before installation, safety 3 Bill of materials 18 C CAUTION, signicance 3 Cleaning the cabinet 13 D Damage, sig.
23 John Bunn O2 Premier Oxygen Concentrator NOTES.
24 John Bunn O2 Premier Oxygen Concentrator REGISTER YOUR PRODUCT! The benets of r egistering: • Safeguard your investment. • Ensure long-term maintenance and ser vicing of your pur chase. • Receive updates with product information, mainte- nance tips, and industr y news.
QUESTIONNAIRE 1. Method of purchase: (check all that apply) � Medicare � Insurance � Medicaid � Other 2. This product was pur chased for use by: (check one) � Self � Parent � Spouse � Other 3. This product was pur chased for use at: � Home � Facility � Other 4.
Affix first-class postage here GF Health Pr oducts, Inc. 2935 Northeast Parkway Atlanta, GA 30360.
.
USA, Corporate Headquar ters: Graham-Field Health Products 2935 Nor theast Parkway Atlanta, Georgia 30360 telephone: 800-347-5678, 770-447-1609 fax: 800-726-0601, 678-291-3232 http://www .
An important point after buying a device Graham Field JB0160-010-220 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Graham Field JB0160-010-220 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Graham Field JB0160-010-220 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Graham Field JB0160-010-220 you will learn all the available features of the product, as well as information on its operation. The information that you get Graham Field JB0160-010-220 will certainly help you make a decision on the purchase.
If you already are a holder of Graham Field JB0160-010-220, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Graham Field JB0160-010-220.
However, one of the most important roles played by the user manual is to help in solving problems with Graham Field JB0160-010-220. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Graham Field JB0160-010-220 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center