The information and technical data disclosed in this document may be used and disseminated only for the purposes and to the extent specifically authorized in writing by General Monitors. Instruction Manual 11/00 General Monitors reserves the right to change published specifications and designs without prior notice.

WARRANTY STATEMENT General Monitors warrants the Model TS420 to be free from defects in workmanship or material under normal use and service within two (2) years (sensor cells one (1) year) from the date of shipment. General Monitors will repair or replace without charge any equipment found to be defective during the warranty period. Full determination of the nature of, and responsibility for, defective or damaged equipment will be made by General Monitors’ personnel. Defective or damaged equipment must be shipped prepaid to General Monitors’ plant or the representative from which shipment was made. In all cases this warranty is limited to the cost of the equipment supplied by General Monitors. The customer will assume all liability for the misuse of this equipment by its employees or other personnel. All warranties are contingent upon proper use in the application for which the product was intended and do not cover products which have been modified or repaired without General Monitors’ approval or which have been subjected to neglect, accident, improper installation or application, or on which the original identification marks have been removed or altered. Except for the express warranty stated above, General Monitors disclaims all warranties with regard to the products sold, including all implied warranties of merchantability and fitness and the express warranties stated herein are in lieu of all obligations or liabilities on the part of General Monitors for damages including, but not limited to, consequential damages arising out of/or in connection with the use or performance of the product. WARNING - THE TS420 DETECTS OXYGEN DEFICIENT ENVIRONMENTS AND EXPOSURE TO OXYGEN DEFICIENT ENVIRONMENTS MAY RESULT IN SICKNESS OR DEATH. The TS420 Toxic Gas Detector contains components, which can be damaged by static electricity. Special care must be taken when wiring the system to ensure that only the connection points are touched. The TS420 is rated intrinsically safe as denoted by the symbol “Ex ia” on the instrument label. Substitution of electrical components within the TS420 may impair intrinsic safety. Damage to the TS420 housing such that any internal components or potting can be or is exposed compromises the intrinsic safety of the device and as such should not be used in a hazardous environment. Such damage includes complete fracture of the plastic housing or cracks that may be opened to expose any internal components or potting. Destruction of the sensor cap will not affect the intrinsic safety of the TS420; however, calibration accuracy and the watertight nature of the TS420 will be compromised. WARNING - DO NOT USE A TS420 WITH A DAMAGED HOUSING IN A HAZARDOUS ENVIRONMENT. IMPORTANT - Each oxygen sensor cell is shipped separate from the TS420. This ensures that a fresh sensor cell will be used during initial start-up. DO NOT install the cell into the TS420 until you are ready to apply power to the system. When installing this unit, an initial field calibration must be completed, since the TS420 is not factory calibrated to the specific cell.

System Integrity Verification General Monitors mission is to benefit society by providing solutions through industry leading safety products, services and systems that save lives and protect capital resources from the dangers of hazardous flames, gases and vapors. The safety products you have purchased should be handled carefully and installed, calibrated and maintained in accordance with the respective product instruction manual. Remember these products are for your safety. To ensure operation at optimum performance General Monitors recommends that certain maintenance items are performed. Commissioning Safety Systems Before power up, verify wiring, terminal connections and stability of mounting for all integral safety equipment including, but not limited to: ! Power supplies ! Control modules ! Field detection devices ! Signaling / output devices ! Accessories connected to field and signaling devices After the initial application of power (and any factory specifies warm-up period) to the safety system, verify that all signal outputs, to and from devices and modules, are within the manufacturers’ specifications. Initial calibration / calibration checking / testing should be performed per the manufacturers’ recommendations and instructions. Proper system operation should be verified by performing a full, functional test of all component devices of the safety system, ensuring that the proper levels of alarming occur. Proper system operation should be verifies by performing a full, functional test of all component devices of the safety system, ensuring that the proper levels of alarming occur. Fault/Malfunction circuit operation should be verified.

Periodic Testing/Calibration of Field Devices Periodic testing/calibrating should be performed per the manufacturers’ recommendations and instructions. Testing/Calibrating procedures should include, but not be limited to: ! Verify zero reading ! Apply a known concentration of gas, or a simulated test device provided by the manufacturer ! Verify integrity of all optical surfaces and devices ! For flame detectors, use the appropriate test lamp When testing produces results outside of the manufacturers’ specifications, re-calibration or repair/replacement of the suspect device(s) should be performed as necessary. Calibration intervals should be independently established through a documented procedure, including a calibration log maintained by plant personnel or third party testing services. Periodic System Verification The following system verifications should be performed at least annually: Verify wiring, terminal connections and stability of mounting for all integral safety equipment including, but not limited to: ! Power supplies ! Control modules ! Field detection devices ! Signaling / output devices ! Accessories connected to field and signaling devices
Proper system operation should be verified by performing a full, functional test of all component devices of the safety system, ensuring that the proper levels of alarming occur. Fault/Malfunction circuit operation should be verified.

1. Mount and Wire the Detector (pages 5 through 9) There are no standard rules for detector placement since the optimum sensor location is different for each application. The customer must evaluate conditions at the facility to make this determination. Generally, the TS420 should be easily accessible for calibration checks. The TS420 should be mounted pointing down to prevent water build-up on the detector head. The detector assembly should not be placed where it may be coated by contaminating substances. Although the Model TS420 is RFI resistant, it should not be mounted in close proximity to radio transmitters or similar equipment. Some general guidelines for sensor location: ! The TS420 should be used in locations that are tightly enclosed with poor or no ventilation, such as utility tunnels or underground storage areas. These areas are likely to have dangerously low oxygen concentrations. ! Observe the TS420’s environmental specifications when locating the unit. The environmental specifications covering the electrochemical cell are found in the Appendix, Section 5.1.4. Detectors should be mounted in an area that is as free from wind, dust, water, shock and vibration as possible. WARNING - Operation above or below temperature limits will provide unstable readings, which may cause false alarms or failure to alarm. See Section 5.1.4 for guidelines. General Monitors discourages painting of detector assemblies. If the detector head is painted over, the gas will not be able to diffuse into the sensor. If the TS420 body is painted over, the LCD display cannot be read. The TS420 is shipped without the sensor installed. The sensor must be installed into the detector and calibrated for proper operation. See Sections 3.3 and 3.4 for installation instructions and Section 4.3 for calibration instructions. Once correctly installed, the TS420 requires little or no maintenance other than periodic calibration checks to ensure system integrity. General Monitors recommends a calibration check schedule be established and that the complete system, including all alarm circuitry, be tested at least annually. The overall and mounting dimensions for the TS420 (Figure Q-A) should be used when making installation determinations. Mechanical specifications can be found in the Appendix, Section 5.1.2.

LOOP- (WHITE)
FRAME GROUND
(GREEN/YELLOW)
1 1/2-20 UN-2A
Figure Q-A Outline and Dimensional Drawing

Intrinsically Safe Barriers In order for the TS420 to be used with classified equipment inside of the hazardous area, it is necessary for the customer to purchase either a Galvanic or Zener Diode Barrier, which should be mounted in the non-classified area between the TS420 and the facilities’ control room. A list of recommended manufacturers and part numbers follows for galvanic barriers: Stahl - 9303/11-22-11 Elcon - mD323 or mD325 MTL - MTL7206 or MTL5041 Turck - mk33-Li-ExO Pepperl+Fuchs - KFD2-CR-EX1.30-300 These are recommended active type barriers for intrinsically safe applications. Each application is different and it is the ultimate decision of the user to determine the appropriate I/S barrier for their application. NOTE - To maintain intrinsic safety, wire length and gauge are limited by the capacitance and inductance of the wire, Ci and Li of the TS420, and the barrier ratings. See Section 5.1.5 for instructions on calculating these values. Align the TS420 so that the LCD display is easily viewed. Teflon tape may be used on the TS420’s threads. The two wires at the base of the TS420 provide a 4-20mA loop-powered output. The red wire is the “loop +” lead and the white wire is the “loop -” lead (Figure Q-B). A third green/yellow wire is provided for frame ground.

Figure Q-B Wiring Diagram
It is recommended that a two wire, shielded cable be used for making power and output signal connections on the TS420. Connect the green/yellow frame ground wire to the grounding terminal or lug inside the junction box being used with the TS420. Ensure the junction box frame is connected to frame ground or connected to the cable shield, which is connected to frame ground at the controller. Connect the red wire to the +24VDC terminal. Connect the white wire to the signal input terminal on the readout/relay display module, FMD Field Mounted Display, S4100E smart sensor, industrial analog to digital converter, computer-based monitor, PLC, DCS, etc. Since the TS420 is designed to operate continuously, a power switch is not included, in order to prevent accidental system shutdown.

NOTE - Power must remain disconnected until all wiring connections are made. The absolute maximum distance between the TS420 and the power supply is 8000 feet (2438 meters). Depending on the application, general purpose or intrinsically safe, the cable length will depend on the intrinsically safe barrier chosen and the cable used. See the Appendix, Section 5.1.5, for the intrinsically safe system requirements. In addition, cable length specifications can also be found in the Appendix, Section 5.1.3. In all cases, the cable run should be as short as possible. 2. Apply power to the detector. Before applying power to the system for the first time, all wiring connections should be checked for correctness. The sensor should be allowed to stabilize prior to calibration. The stabilization time should be one (1) hour for all sensors. Initial Power-up The TS420 has been shipped un-calibrated, therefore, the first time power is applied, the TS420 will enter Start-up mode (3.6mA) for a short interval, then may transition to fault mode (flashing MODE LCD with the CAL LCD off and the analog output current at 3.5mA) until calibrated (See Section 4.3 for Calibration Procedure). After calibration is complete, the TS420 will proceed to Operate mode (17.38mA = 20.9% V/V O2) where any detected oxygen will show as a proportional analog output current between 4 and 20mA (4mA = 0% V/V O2, 20mA = 25% V/V O2). The MODE LCD will be on steady and the CAL LCD off with normal atmosphere and both the MODE and CAL LCD's will flash when oxygen deficiency is detected. (O2 level < 19%). General Monitors recommends the TS420 be calibrated again within the first twenty-four (24) hours after the initial calibration. Calibrations should be performed with new units and units that have been off power for more than one (1) week. See “Calibration”, Section 4.3, for more details. Power-up After Initial Calibration When the TS420 is powered up, after it has been previously calibrated, the unit will enter a Start-up mode (3.6mA) for approximately 50 seconds, which allows the sensor to stabilize before proceeding to the Operate mode (17.38mA). During Start-up, the “MODE” LCD will flash, the “CAL” LCD will be off and the analog output current will be 3.6mA.
Figure Q-C Operating Modes, Indicators and Outputs

3. Calibrate the Detector. Activating the “CAL” switch will automatically disable the external alarm circuits by fixing the analog output current to 3.75mA. General Monitors recommends that the TS420 be calibrated within the first twenty-four (24) hours after initial start-up, and that calibration be checked at least every ninety (90) days to ensure the integrity of the system. NOTE: A calibration check consists of applying a 20.9% V/V O2 (“normal” ambient air) of full- scale concentration of gas to the sensor and observing the reading on the display device being used. General Monitors is not implying that the customer should expect problems with sensor life or stability. “Frequent” calibration checks merely ensure the integrity of the life protecting equipment, and are recommended for problem environments (i.e. sensors accidentally being painted over, etc, see Appendix 5.4). A calibration schedule should be established and followed. A logbook should also be kept showing calibration dates and dates of sensor replacement. Calibration Procedure: 1. If it is suspected that the air is not at “normal” levels (20.9% O2), it will be necessary to calibrate the sensor with clean air. The 20.9% O2 gas cylinder (GMI P/N 1400262-11) can be used. Apply the clean air source prior to entering calibration mode and allow the sensor to stabilize for at least one (1) minute before proceeding. 2. Place the magnet over the GMI Logo on the body of the unit and hold it there until the “CAL” LCD indicator appears, then remove the magnet. The analog output current will drop to 3.75mA. 3. The “CAL” and “MODE” LCD indicator will flash, indicating that the unit is now calibrating. 4. After approximately 30 seconds, the display will change from a flashing “CAL” and “MODE” LCD to a solid “MODE” LCD, indicating that the calibration is complete and the unit has re- entered “OPERATE” mode. Remove any calibration gas. 6. The unit is now calibrated and the new values have been stored in the NOVRAM (non- volatile memory). Figure Q-D shows a flow diagram of the codes that will appear in the display window during the calibration procedure. If there is a problem and the TS420 cannot complete the calibration sequence, a fault indicator will be displayed, and the analog output current will drop to 3.5mA.
Figure Q-D Flow Diagram

NOTE - If the unit fails to calibrate a flashing “MODE” LCD will be displayed. The sensor must see “clean” air for at least fifteen (15) minutes before a second calibration is attempted by reapplying the magnet. The instrument is now ready to operate! Please consult the manual for more information on the instrument’s many features. If you have any problems in the set-up or testing of the detector, please refer to the “Trouble Shooting Section”, or call the factory direct. Worldwide Service is available by calling: Lake Forest, California Phone: +1-949-581-4464 (24 hr. service) Fax: +1-949-581-1151





1.0 Introduction 1.1 General Description The Model TS420 is a 4-20mA, loop-powered Oxygen Deficiency detector used for monitoring Oxygen levels. The micro-processor based electronics process information at the sensor site and are contained within a corrosion and solvent resistant plastic housing. An LCD display provides status indications that can be viewed through a window in the housing. The analog signal (0 to 25% V/V O2 full scale) provides remote and/or discrete indications of the sensor’s operation. The TS420 is entity-approved as intrinsically safe for use in Class I, Divisions 1 & 2, Groups A, B, C and D, Class II, Groups E, F and G & Class III hazardous areas. It can also be used in general-purpose (non-hazardous) applications.
NOTE - Intrinsically safe applications require an I/S barrier mounted in a safe area. See Section 3.3.1 for guidelines. 1.2 Features and Benefits Microprocessor-Based Electronics: monitors for fault conditions, processes input signals from the sensor and provides outputs in the form of display codes and an analog signal. One Person Adjustment-Free Calibration: initiate the calibration sequence with a magnet, apply 20.9% V/V O2, and wait for the display to indicate that the unit has completed the calibration. No user adjustments are required. Two Segment LCD: indicates gas presence, operational modes, fault codes and calibration cues. 4 to 20 mA Output: transmits fault, calibration, and gas concentration levels to a remote display, computer or other device.
1.3 Applications This is a partial listing of applications. ! Wastewater Treatment Plants ! Chemical and Petrochemical Plants ! Refineries ! Pulp & Paper Plants ! Micro-chip Manufacturing ! Pharmaceuticals ! Food & Beverage ! Utilities

Figure 1 Model TS420

This page left intentionally blank

2.0 Detector Assembly 2.1 Sensor Assembly General Monitors uses a two-electrode electrochemical cell (Figure 2) to provide a stable and accurate gas detector.

Figure 2 Oxygen Deficiency Sensor Assembly
The oxygen sensor is of the self-powered, diffusion limited, metal air battery type. It contains an anode, electrolyte and an air cathode to which the diffusion of oxygen is severely restricted by a diffusion barrier. At the air cathode, oxygen is reduced to hydroxyl ions, which in turn oxidize the metal anode. This action produces a current that is proportional to the rate of consumption of oxygen.

2.2 Control Electronics The Model TS420’s electronics are contained within a plastic housing so that sensor information can be processed at the point of detection. The Model TS420 provides a 4-20 mA output signal proportional to 0 to 25% V/V O2 full-scale gas concentration at the detector. Ambient Air is approximate 20.9% V/V O2. This level equals 17.38 mA. As the oxygen level decreases the 4-20 mA signal decreases. The TS420 provides a two-digit LCD (Liquid Crystal Display). This display (Figure 3) provides status such as gas present, fault and calibration cues to the operator. The MODE LCD is on steady during normal operation to show that the sensor is functioning properly. Monitored faults are: data memory failure, failed to calibrate and sensor removed. Engineering specifications covering the sensor cell and control electronics are found in the Appendix, Section 5.1.

Figure 3 LCD Display Location

3.0 Installation 3.1 Receipt of Equipment All equipment shipped by General Monitors is packed in shock absorbing containers, which provide considerable protection against physical damage. The contents should be carefully removed and checked against the packing slip. If any damage has occurred or if there is any discrepancy in the order, notify General Monitors as soon as possible. All subsequent correspondence with General Monitors must specify the equipment part number and serial number. Each TS420 is thoroughly checked by the factory. However, a complete check-out and initial calibration is necessary upon initial installation and start up to ensure system integrity. 3.2 Detector Location Considerations There are no standard rules for detector placement since the optimum sensor location is different for each application. The customer must evaluate conditions at the facility to make this determination. Generally, the TS420 should be easily accessible for calibration checks. The TS420 should be mounted pointing down to prevent water build-up on the detector head. The detector assembly should not be placed where it may be coated by contaminating substances. Although the Model TS420 is RFI resistant, it should not be mounted in close proximity to radio transmitters or similar equipment. Some general guidelines for sensor location: ! The TS420 should be used in locations that are tightly enclosed with poor or no ventilation, such as utility tunnels or underground storage areas. These areas are likely to have dangerously low oxygen concentrations. ! Observe the TS420’s environmental specifications when locating the unit. The environmental specifications covering the electrochemical cell are found in the Appendix, Section 5.1.4. Detectors should be mounted in an area that is as free from wind, dust, water, shock and vibration as possible. WARNING - Operation above or below temperature limits will provide unstable readings, which may cause false alarms or failure to alarm. See Section 5.1.4 for guidelines. General Monitors discourages painting of detector assemblies. If the detector head is painted over, the gas will not be able to diffuse into the sensor. If the TS420 body is painted over, the LCD display cannot be read.

3.3 Installation Instructions The TS420 is shipped without the sensor installed. The sensor must be installed into the detector and calibrated for proper operation. See Section 4.5 for installation instructions and Section 4.3 for calibration instructions. Once correctly installed, the TS420 requires little or no maintenance other than periodic calibration checks to ensure system integrity. General Monitors recommends a calibration check schedule be established and that the complete system, including all alarm circuitry, be tested at least annually. The overall and mounting dimensions for the TS420 (Figure 12) should be used when making installation determinations. Mechanical specifications can be found in the Appendix, Section 5.1.2. 3.3.1 Intrinsically Safe Barriers In order for the TS420 to be used with classified equipment inside of the hazardous area, it is necessary for the customer to purchase either a Galvanic or Zener Diode Barrier, which should be mounted in the non-classified area between the TS420 and the facilities’ control room. A list of recommended manufacturers and part numbers follows for galvanic barriers: Stahl - 9303/11-22-11 Elcon - mD323 or mD325 MTL - MTL7206 or MTL5041 Turck - mk33-Li-ExO Pepperl+Fuchs - KFD2-CR-EX1.30-300 These are recommended active type barriers for intrinsically safe applications. Each application is different and it is the ultimate decision of the user to determine the appropriate I/S barrier for their application. NOTE - To maintain intrinsic safety, wire length and gauge are limited by the capacitance and inductance of the wire, Ci and Li of the TS420, and the barrier ratings. See Section 5.1.5 for instructions on calculating these values. Align the TS420 so that the LCD display is easily viewed. Teflon tape may be used on the TS420’s threads. 3.3.2 Maintenance The removal of particulate matter from sensor accessories may be done through the use of clean water only. Solvents should not be used. The accessories should be thoroughly dried, with compressed air if necessary, before refitting to the sensor body. Do not apply compressed air directly to the sensor. Some typical items to check during maintenance examinations are: ! The sensor mounting, to see it is secure. ! The sensor, to see it is clear of oil, water, dust or paint which might clog it. ! The cable connections for tightness and possible damage. ! All detector placements are up-to-date with the layout of the plant, e.g. modifications to the plant. ! The complete system, to see it has a back up supply for the full-prescribed time.

3.4 Wiring Connections The two wires at the base of the TS420 provide a 4-20mA loop-powered output. The red wire is the “loop +” lead and the white wire is the “loop -” lead (Figure 4). A third green/yellow wire is provided for frame ground. It is recommended that a two wire, shielded cable be used for making power and output signal connections on the TS420. Connect the green/yellow frame ground wire to the grounding terminal or lug inside the junction box being used with the TS420. Ensure the junction box frame is connected to frame ground or connected to the cable shield, which is connected to frame ground at the controller. Connect the red wire to the +24VDC terminal. Connect the white wire to the signal input terminal on the readout/relay display module, FMD Field Mounted Display, S4100E smart sensor, industrial analog to digital converter, computer-based monitor, PLC, DCS, etc. Since the TS420 is designed to operate continuously, a power switch is not included, in order to prevent accidental system shutdown. NOTE - Power must remain disconnected until all wiring connections are made. The absolute maximum distance between the TS420 and the power supply is 8000 feet (2438 meters). Depending on the application, general purpose or intrinsically safe, the cable length will depend on the intrinsically safe barrier chosen and the cable used. See the Appendix, Section 5.1.5, for the intrinsically safe system requirements. In addition, cable length specifications can also be found in the Appendix, Section 5.1.3. In all cases, the cable run should be as short as possible.

Figure 4 Wiring Diagram

4.0 Operation 4.1 Applying Power Before applying power to the system for the first time, all wiring connections should be checked for correctness. The sensor should be allowed to stabilize prior to calibration. The stabilization time should be one (1) hour for all sensors. Initial Power-up The TS420 has been shipped un-calibrated, therefore, the first time power is applied, the TS420 will enter Start-up mode (3.6mA) for a short interval, then may transition to fault mode (flashing MODE LCD with the CAL LCD off and the analog output current at 3.5mA) until calibrated (See Section 4.3 for Calibration Procedure). After calibration is complete, the TS420 will proceed to Operate mode (17.38mA = 20.9% V/V O2) where any detected oxygen will show as a proportional analog output current between 4 and 20mA (4mA = 0% V/V O2 , 20mA = 25% V/V O2). The MODE LCD will be on steady and the CAL LCD off with normal atmosphere and both the MODE and CAL LCD's will flash when oxygen deficiency is detected. (O2 level < 19%). General Monitors recommends the TS420 be calibrated again within the first twenty-four (24) hours after the initial calibration. Calibrations should be performed with new units and units that have been off power for more than one (1) week. See “Calibration”, Section 4.3, for more details. Power-up After Initial Calibration When the TS420 is powered up, after it has been previously calibrated, the unit will enter a Start-up mode (3.6mA) for approximately 50 seconds, which allows the sensor to stabilize before proceeding to the Operate mode (17.38mA). During Start-up, the MODE LCD will flash, the CAL LCD will be off and the analog output current will be 3.6mA.
OPERATING MODES INDICATOR ANALOG OUTPUT CAL MODE START-UP ❏ ✹ 3.6mA OPERATE AMBIENT O2 = 20.9% ❏ ■ 17.38mA DEFICIENT ATMOSPHERE, O2< 19%1 ✹ ✹ 4.0-16.16mA FAULT NO POWER ❏ ❏ 0.0mA SYSTEM ❏ ✹ 3.5mA
Figure 5 Operating Modes, Indicators and Outputs

4.2 Operating Modes The TS420 has four distinct operating modes: Start-up, Normal Operation, Calibration and Fault. The mode is indicated by both the analog output and the LCD display. See Figures 5 and 6.
4.3 Calibration Activating the “CAL” switch will automatically disable the external alarm circuits by fixing the analog output current to 3.75mA. General Monitors recommends that the TS420 be calibrated within the first twenty-four (24) hours after initial start-up, and that calibration be checked at least every ninety (90) days to ensure the integrity of the system. NOTE: A calibration check consists of applying a 20.9% V/V O2 (“normal” ambient air) of full- scale concentration of gas to the sensor and observing the reading on the display device being used. General Monitors is not implying that the customer should expect problems with sensor life or stability. “Frequent” calibration checks merely ensure the integrity of the life protecting equipment, and are recommended for problem environments (i.e. sensors accidentally being painted over, etc, see Appendix 5.4). A calibration schedule should be established and followed. A logbook should also be kept showing calibration dates and dates of sensor replacement. Calibration Procedure: 1. If it is suspected that the air is not at “normal” levels (20.9% O2), it will be necessary to calibrate the sensor with clean air. The 20.9% O2 gas cylinder (GMI P/N 1400262-11) can be used. Apply the clean air source prior to entering calibration mode and allow the sensor to stabilize for at least one (1) minute before proceeding. 2. Place the magnet over the GMI Logo on the body of the unit and hold it there until the “CAL” LCD indicator appears, then remove the magnet. The analog output current will drop to 3.75mA. 4. The “CAL” and “MODE” LCD indicator will flash, indicating that the unit is now calibrating. 4. After approximately 30 seconds, the display will change from a flashing “CAL” and “MODE” LCD to a solid “MODE” LCD, indicating that the calibration is complete and the unit has re- entered “OPERATE” mode. Remove any calibration gas. 6. The unit is now calibrated and the new values have been stored in the NOVRAM (non- volatile memory). Figure 6 shows a flow diagram of the codes that will appear in the display window during the calibration procedure. If there is a problem and the TS420 cannot complete the calibration sequence, a fault indicator will be displayed, and the analog output current will drop to 3.5mA.


Figure 6 Flow Diagram
4.4 Calibration Equipment General Monitors offers flow calibration kits as the method of introducing calibration gas to the TS420. (See Section 5.5). Calibration is performed with 20.9% V/V O2. NOTE - When removing regulator from gas cylinders, open value on regulator to ensure that gas is purged and pressure in the regulator is relieved. Be sure to follow this recommendation when switching cylinders. WARNING – Do not use a gas cylinder if the regulator gage shows the cylinder pressure in the black area just above 0 psi. Unstable calibration could result.
If the TS420 is to be used with the optional Splash-Guard, all calibrations must be performed using the calibration cup (Figure 7) with the Splash-Guard installed onto the TS420. In addition, all gas checks must be performed using the calibration cup with the Splash-Guard installed. This performs two functions: 1. It provides a more accurate representation of the amount of gas getting to the sensor. 2. It checks that the Splash-Guard is not clogged and that gas can get to the sensor.

Figure 7 Calibration Cup
If the Splash-Guard will not be used, the TS420 must be calibrated and gas checked with the calibration plug (Figure 8)
Figure 8 Calibration Plug


0 10 20 30 40 50 60 70 80 90 100
Figure 9 Change in % Oxygen in the atmosphere with Relative Humidity at constant Temperature
4.5 Sensor Replacement Procedure The following is the procedure to replace a TS420 sensor.
% Oxygen · Relative Humidity (%)



4.6 Sensor Care and Maintenance If power is removed from the TS420 for periods greater than one hour, the sensor should be removed from the TS420 and placed in the original storage container shipped with the sensor. Storage of the sensor in the original container guards against sensor contamination from solvents, lubricant humidity, etc. NOTE - Do not short the sensor pins.
In the case of long intervals of non-use, the sensors should be stored as described above in a cool, dry place, preferably refrigerated between 0 and 20°C.
In all cases, after the sensor has been removed from the TS420, the cap should be replaced on to the TS420 body and the gas port taped off to prevent any corrosion damage to the exposed socket pins.
The TS420 has self-diagnostics incorporated into the microprocessor program. If a fault is detected, the output signal will drop to 3.5mA and the MODE LCD will be flashing to indicate the fault. When a calibration fault has occurred and has been corrected, the unit will stay in fault until it has been successfully calibrated. The Fault Codes are: Fault (No Power): Neither LCD will be on. Output signal will be 0.0mA. Remedy: Check wiring, ensure power is applied to unit. Fault (System): The “MODE” LCD will be flashing. Output signal will be 3.5mA. Remedy: Attempt recalibrating the detector. If the fault persists, replace the sensor and after sensor has stabilized (one (1) hour minimum). Recalibrate. If fault still persists, contact factory or your sales representative.

5.1.1 System Specifications Sensor Type: Electrochemical cell Typical Life (sensing element): 2 years in normal air Warranty: Two years for the electronics One year for the sensor cell Malfunctions Monitored: Calibration Errors Data Memory Errors Measuring Range: Oxygen 0-25% V/V O2 Response Time: T90 < 15 sec Repeatability: + 1 % full scale Approvals: UL, CUL, CSA, CENELEC and CE approved.
5.1.3 Electrical Specifications Recommend two wire 14, 16 or 18 AWG shielded cable.
General Purpose Installations Maximum distance between the TS420 and the power source @ 24VDC nominal (600 Ω load resistor maximum) is 8000 feet (2438 meters).
Intrinsically Safe Installations See Sections 5.1.5 through 5.1.7.

Intrinsically Safe Parameters Ui /VMAX = 35VDC Ii /IMAX = 100mA Ci = 0 Li = 500 µH Input Power: 10 to 35VDC range; 3.5 to 22mA loop powered Output Current: (600 Ω max. @ 24VDC) (120 Ω max. @ 10VDC) Signal Range 3.5 to 22mA Fault 3.5mA Start-up 3.6mA Calibration 3.75mA Detection Range 4 to 20mA Over-range 22mA
Electrical Classification: Class I, Divisions 1 & 2, Groups A, B, C & D; Class I, Zone 0, AEx ia IIC T6; EX ia IIC T6; Eex ia IIC T6 Class II, Groups E, F & G, Class III, IP67, Type 4X
RFI/EMI Protection: Complies with EN50081-2, EN50082-2 Status Indicator: Liquid Crystal Display with Normal, Fault, and Calibration cues 5.1.4 Environmental Specifications Operating Temperature Range: -4°F to +122°F (-20°C to +50°C) Storage Temperature Range: -4°F to +122°F (-20°C to +50°C) Humidity Range: 0 to 99% relative humidity non-condensing Pressure Range: Atmosphere ± 10%

5.1.5 Intrinsically Safe System Requirements NOTE - Refer to control drawing 45217 Rev A. (Section 5.2.2) The TS420 is approved as intrinsically safe on its own (entity approved). However, when installing the TS420 into a system, there are requirements, which must be strictly followed to maintain the intrinsic safety of the system. These requirements are:
Wire Inductance/Ft 5.1.6 Sample Calculation What is the maximum allowable cable length when using an MTL 7206 barrier and BELDEN 8760 2-conductor shielded cable? BELDEN 8760 Parameters: C/ft = 24 pF/ft, L/ft: not available MTL 7206: TS420: CMAX = 0.12 µf Ci = 0 µf LMAX = 4.0 mH Li = 500 µH Ui/VMAX = 35V Ui/VMAX = 35V Ii/IMAX = 93mA Ii/IMAX = 100mA 1. Ui/VMAX (TS420) ≤ Ui/VMAX (MTL 7206) 35VDC ≤ 35VDC 2. Ii/IMAX (TS420) ≥ Ii/IMAX (MTL 7206) 100mA 93mA 3. Cable inductance not available. 4. Max. cable length = CMAX (MTL 7206) - Ci (TS420) C/L (Belden 8760) = 0.12µf - 0µf 24pF/ft Max. Cable length = 5000 feet

5.2.1 Outline & Dimensional Drawing

Figure 12 Outline and Dimensional Drawing

5.2.2 Control Drawing

Figure 13 Control Drawing


MAX CABLE LENGTH = MAX BARRIER CAPACITANCE (Cmax) - INTERNAL TS420
MAX CABLE LENGTH BASED ON CAPACITANCE= (0.13uF - 0uF) / (24pF/ft) = 5416ft
MIN TS420 VOLTAGE < FIELD DEVICE VOLTAGE AT MAX CURRENT
10V < 28V - ((22mA)370 +1.5V) - ((22mA)50 + 2.1V) - (22mA)30 = 14.5V
Figure 14 Sample Installation

5.2.4 Junction Boxes

Figure 15 10252 Round Aluminum Junction Box

Figure 16 31305-2 Aluminum Junction Box


Figure 17 45160-1 Plastic Junction Box, NPT; 45160-2 Plastic Junction Box, M20
5.2.5 Splash-Guard

Figure 18 70631-2 Splash - Guard

5.2.6 Accessories

Figure 19 45170-1 Flow Block Figure 20 45147-1 Flow Block Retainer
5.2.7 Calibration Accessories
Figure 21 914-152 Calibration Plug

Figure 22 1400152-1 Calibration Cup

5.3 Ordering Information TS420 - 3 0 0 - 101 - 1 0
Junction Box: 0 = None 1 = 45160-1 Junction Box – Plastic NPT 3 = 10252 Junction Box – Round, Aluminum

5.4 Calibration Schedule for Problem Environments Detector Serial Number: Location: 1) Installation and preliminary calibration. Record date after preliminary calibration is performed: Date: 2) 24-hour calibration. Record date after 24-hour calibration is performed: Date: 3) 7 day calibration check (Record date and reading of calibration check. Repeat after 7 days if reading deviates more than ±10%. Otherwise go to step 4). Date Reading Date Reading Date Reading 4) 14 day calibration check (Record date and reading of calibration check. Repeat after 14 days if reading deviates more than ±10%. Otherwise go to step 5). Date Reading Date Reading Date Reading 5) 30 day calibration check (Record date and reading of calibration check. Repeat after 30 days if reading deviates more than ±10%. Otherwise go to step 6). Date Reading Date Reading Date Reading 6) 60 day calibration check (Record date and reading of calibration check. Repeat after 60 days if reading deviates more than ±10%. Otherwise go to step 7). Date Reading Date Reading Date Reading 7) 90 day calibration check: Date Reading Date Reading Date Reading

5.5 Spare Parts & Accessories TS420 Spare Parts The Model TS420 has potted electronics, therefore, there are no replacement boards available for this unit. The part numbers of the TS420 itself are: TS420 Oxygen Deficiency 45200-1 Sensor Cap 45103-1 Replacement Sensors Oxygen Deficiency Sensor 45213-1 TS420 Mounting Accessories Splash-Guard (Standard) 70631-2 Flow Block (Incl. Ret./Cal Plug) 45170-1 Retainer, Flow Block 45147-1 Alum. Junction Box (Small) 10252 Alum. Junction Box (Large) 31305-2 Plastic Junction Box (3/4” NPT) 45160-1 Plastic Junction Box (M20) 45160-2 Cable Grip 961-008 ¾” NPT, Adapter, Plastic 961-009 ¾” NPT Adapter, Steel 961-007 20 mm x ¾” NPT Adapter, Brass 961-006 TS420 Calibration Accessories Tubing 931-085 Regulator (500ml/Min) 922-023 Calibration Plug 914-152 Calibration Cup 1400152-1 Calibration Kits (cylinder, regulator & tubing) O2 20.9% 1400263-11 Spare Cylinders O2 20.9% 1400262-11