Druck DBC 150/650 Series
Dry block temperature calibrator
User manual - K400
[EN] English 1 56 [FR] French TBA [DE] German TBA [IT] Italian TBA [PT] Portuguese TBA [ES] Spanish TBA


© 2005 General Electric Company. All rights reserved.

Technical Advice • For technical advice contact the manufacturer.
*A qualified technician must have the necessary technical knowledge, documentation, special test equipment and tools to carry out the required work on this equipment.
This equipment meets the requirements of all relevant European safety directives. The equipment carries the CE mark.
K400 Issue No. 1 i

TRACEABILITY AND CALIBRATION REPORT • Each DBC instrument is supplied as standard with a calibration report, declaring traceability to a National or International standard.
1 Our equipment is guaranteed against defective material and workmanship during the warranty period. Claims under warranty can be made by returning the equipment, prepaid, to our factory. The equipment will be replaced, repaired or adjusted at our discretion.
2 The liability of GE is restricted to that given under our warranty. No responsibility is accepted for damage, loss or other expense incurred through the sale or use of our equipment.
3 Under no circumstances will GE be liable for any special incidental or consequential damage.
The following symbols are used to identify hazards on the instrument:
This symbol, on the instrument, indicates that the user should refer to the user manual.
This symbol, on the instrument, indicates a very hot surface that must not be touched.
K400 Issue No. 1 ii

Safety Instructions This instrument should only be used by suitably qualified * technicians
During operation observe the following safety instructions:
1 USE the instrument in vertical position only.
2 DO NOT place any object on the protection grid of the dry well.
3 DO NOT put any liquid into the dry well and/or on the inserts.
4 DO NOT use fluids to clean the dry well and/or the inserts.
5 BEFORE any maintenance activity, WAIT for the instrument to cool to ambient temperature (keep the instrument switched on, to allow the cooling fan to operate).
6 DO NOT touch the protection grid until it has cooled to ambient temperature).
7 The instrument must be at ambient temperature before storing in any transit case or packing system.
8 DO NOT use this instrument near inflammable liquids or in Hazardous areas.
9 DO NOT place either your hands, or any other object, into the dry well.
10 Before use, REMOVE all traces of contamination from the device under test.
11 During normal operation, this instrument could reach high temperatures.
12 Always observe all the safety instructions.
*A qualified technician must have the necessary technical knowledge, documentation, special test equipment and tools to carry out the required work on this equipment.
The following abbreviations are used in this manual; the abbreviations are the same in the singular and plural.
BLK block ent enter CJ Cold Junction CJC Cold Junction Compensation DBC Dry Block Calibrator °C Degrees Centigrade in inch mA milliampere mm millimetre mV millivolt RDG reading RTD Resistive Temperature Device T/C Thermocouple V Volt TC Temperature Calibrator TS Temperature Source
K400 Issue No. 1 iii







The DBC-TC series is supplied with the following standard accessories:
2 test leads and 2 clamps for switch connection
6 test leads and 6 clamps for connection of electrical signal
K400 Issue No. 1 vii

The DBC-TC series is supplied with the following standard accessories:
2 test leads and 2 clamps for switch connection
K400 Issue No. 1 viii

The DBC series of temperature calibrators test and calibrate temperature devices. The series comprises two types: a TS (Temperature Source) and a TC (Temperature Calibrator). Multilingual firmware provides a menu-driven user interface in English, French, German, Italian, Spanish and Portuguese. An RS232 digital interface, supplied as standard, enables remote operation of the instrument from a compatible computer.

reference) and a possible electrical feedback from a device under test (i. e. temperature probes, temperature transmitters). An error calculation, with scaling function, allows the user to check immediately the accuracy specification of the unit under test. A 24 V auxiliary excitation provides power for temperature transmitters under test. 1.1 Automatic Calibration


Figure 1-1

BEFORE OPERATING, CHECK THAT THE POWER SUPPLY SELECTOR SWITCH SETTING MATCHES THE POWER SUPPLY VOLTAGE.
The DBC series of calibrators operate from an a. c. power supply of 100 to 120 Vac or 220 to 240 Vac, 50-60 Hz. CAUTIONS:
1 THE POWER SUPPLY MUST PROVIDE CONNECTION TO A PROTECTIVE GROUND TERMINAL. 2 THE INSTRUMENT MUST, AT ALL TIMES, BE CONNECTED TO THE SUPPLY EARTH (GROUND). 2.3 Temperature Probes WARNING
AFTER USE, THE PROBES WILL BE VERY HOT. HANDLE WITH CARE AND WAIT FOR THE PROBES TO COOL COMPLETELY.
The temperature probe to be tested must be correctly placed in the dry well to produce accurate test results:

YES
NO
NO YES
YES
* dimension should comply with EA-10/13 guidelines - for temperature ranges -80° C to +660°C maximum 0.5mm IMPORTANT NOTE
Reduction inserts with special holes could be necessary when carrying out a calibration procedure on temperature device with specific mechanical dimensions. Refer to GE for details of specific application solutions.











Calibration with external reference probe The temperature value to be considered as standard is measured with an external and certified reference probe. This measurement reading can be directly shown either on the DBC display (connecting the reference probe to the REF RTD socket on the top of the instrument) or with an external, certified electronic calibrator.


Power, reading and calibration of devices under test (DBC-TC only) The DBC-TC series has the input capability to directly measure the electrical signal of unit under test (T/C, RTD, mA, V, Ohms): using an error calculation function (with scaling capability), calibration procedures can be carried out and the UUT accuracy immediately measured. A 24 V auxiliary excitation supply is also available to power the loop of the device due for calibration. The generated temperature can be measured with the internal or the external reference as described above.


1 Wire 1 2 Wire 1a 3 Wire 2 4 Wire 2a

5 Pressing this key recalls a related display containing more information, either on the selected function or on the operation to be performed. Pressing this key after the # key, changes the display containing information relating to the selected temperature reference or to the selected test function (for example switch test).

6 Pressing this key changes the sign on working screen. Pressing this key after the # key, recalls the Switch Test function. Pressing # and +/- again to return to the working screen.
7 Pressing this key sets the decimal point. Pressing this key after the # key, recalls the step generation mode, starting from the step selection screen. • 0 key
8 Pressing this key zeroes the display. Pressing this key after the # key, recalls the ramp generation mode, starting from the Ramp set-up screen. • 1- 9 (numeric) keys
9 Using these numeric keys sets a temperature value or a special generation mode (ramp limits, programmable steps etc.). Pressing keys 1- 5 after the # key, recalls specific preset values.
10 Pressing these keys moves the screen cursor either in the horizontal or vertical direction. When the ramp and step functions are enabled, these keys can start, stop, hold or recall related values.

The DBC-TC series has four operating displays:
REF T.
INTERNAL SEL RTD
NONE
NONE
INTERNAL SEL RTD
MEASURE
NONE
DBC-TS
This display shows the measure parameters. The arrow keys move the cursor in both horizontal and vertical directions.
LANGUAGE TEMPERATURE UNIT KEYBOARD SERIAL PORT PRINT
Use the keys to move the cursor up/down and press ent to select.
T/ C K 23.8° C
RDG: °C 0.15% SPAN 148.50
SWITCH
150.00 DIRECT
BLK °C 148.00
150.00 DIRECT
BLK °C 148.00
T/C RTD mA · T/C RTD mA

This display shows the temperature set-point, the actual temperature (with internal or external reference) and the current temperature measurement. Use the numeric key-pad to set the temperature value and, using the second function key, change the direct function to ramping, stepping and preset.
INTERNAL SEL RTD
NONE
SELECT ENT TO CONFIRM
This display shows more information relating to the selected function or test/ measure - (for example selected temperature reference – internal or external -, switch test, mA reading etc.).
V x. xx - mm/ yyyy
Self test
Eeprom OK Temp. Control OK

REF T. 150.00 DIRECT
MEASURE
BLK
INTERNAL SEL RTD
NONE
148.00 °C
Press the MENU key to recall the Menu Selection Screen from any Working Screen. Leave the MEASURE (Input) display set to NONE to display the dry well (BLK) temperature with the temperature set-point shown in small numbers and arrow symbols indicating either a heating or cooling operation.
MEASURE
RDG: °C 0.15% SPAN 148.50
INTERNAL SEL RTD
NONE
150.00 DIRECT
T/C
BLK °C 148.00
From the Menu Selection Screen, use the horizontal and vertical cursor keys move the cursor through the REF. T and MEASURE displays. Select both the reference for the temperature to be generated and an input option (measure). The final working screen, configured as a dual display, shows two sections of equal size, dry well temperature and the unit under test feedback (with or without scaling capability).
T/C RTD mA · RTD mA

The DBC series measures the dry well temperature in two different ways:
REF T.
NONE
INTERNAL SEL RTD
T/C
Using the keys, move the selection cursor on the REF. T display. Select INTERNAL with the keys and press ent to activate the dry well temperature measure made with the internal regulation probe. The next display allows the setting-up of the temperature generation mode (Direct, Ramp, Step or Preset).
REF T.
NONE
INTERNAL SEL RTD
The next selection display now appears as follows.
RTD mA · T/C RTD mA

Move the cursor with the keys, to select a REF 1or REF 2 connection and press ent.
REF 1 REF 2
The following display allows the set-up of the temperature generation mode. Using this facility, the user can connect one of two certified probes to the Ref RTD connector and display the measured dry well temperature.

3.5 System Set-up • Language setting The DBC series have multi-lingual firmware with a default setting of English. To select a different language, press the # + menu keys to recall the System Set-up Screen. Using the keys select the required LANGUAGE (English, French, German, Italian, Portuguese or Spanish) and press ent. Press ce to return to the Working Screen.
LANGUAGE
LANGUAGE TEMPERATURE UNIT KEYBOARD SERIAL PORT PRINT
ENGLISH FRENCH GERMAN
PORTUGUESE ITALIAN
TEMPERATURE UNIT
LANGUAGE
°C
TEMPERATURE UNIT
°F K
KEYBOARD SERIAL PORT PRINT

keys select KEYBOARD and press ent. Select the required function (beep off/on) and press ent again. Press ce to return to the Working Screen.
KEYBOARD
LANGUAGE TEMPERATURE UNIT
BEEP OFF BEEP ON
KEYBOARD
SERIAL PORT PRINT
SERIAL PORT
LANG UAGE TEMPERATURE UNIT KEYBOARD
To set the RS232 communication address, select the System Set-up Screen, using the keys select SERIAL PORT and press ent. To set the communication address, use the numeric key-pad and press ent. Press ce twice to exit the System Set-up Screen.

PRINT SETUP
LANGUAGE TEMPERATURE UNIT KEYBOARD SERIAL PORT PRINT
Period: 0 sec SET NO BLK NO REF NO RDG NO.
CALIBRATION ENTER CODE
ENTER CODE

CALIBRATION ENTER CODE
ENTER CODE
ENTER CODE
4321 CODE:
ENTER CODE
Note:
This procedure permanently changes the PIN code. Using the numeric key-pad, set the new code and press ent to store it.

3.6 Testing Modes • Direct mode After switching on and self-test the display automatically changes to the DIRECT generation working screen. The related display shows the actual dry well temperature in large characters (from the internal or external probe), temperature set-point and a possible measurand (i.e. the last selections made).
BLK
148.00 °C
The status bar shows the temperature set-point, the symbol (indicating a cooling or heating condition) and the generation mode. To set a new set-point value and generate a new temperature, use the numeric key-pad and press ent. The calibrator immediately activates the temperature regulator and the status bar shows the new set-point, with the symbol indicating the new heating or cooling condition. On reaching the new set-point, the RDY (ready) message replaces the
symbol.
BLK
150.00 °C

PROGRAM 20% 25% 33% 50%
Using the keys, select the required percentage of temperature interval between each step or, depending on the application, program a variety of values.
PROGRAM
Using the keys, select one of the standard percentage steps (20, 25, 33, and 50%) and press ent. The following display allows the temperature range span to be set-up (e. g.), 0 to 100 %.
0% = 0.00 °C 100% = 100.00 °C
Using the keys, move the selection cursor in vertical direction, set the required temperature values (0 and 100 %) and press ent. In the next working screen, use the keys to increase or decrease the temperature set-point with the selected percentage difference between each step.
20% · 25% 33% 50%

The status bar shows the recalled percentage step, together with the related temperature set-point, the step symbol ( ) and the heating/cooling arrow symbols.
20.00 20%
19.00 °C
PROGRAM
Using the keys select the number of step to be set and press ent. In the next display, move the selection cursor using the again and set a specific temperature value for each step.
PROGRAMMED STEPS
2 3 4 5
# 1 = 0.00 °C # 2 = 100.00 °C
At the end of this operation, press ent to activate the standard working screen. Use the keys to increase/decrease the step value as described above for the fixed step capability.
20% · 25% 33% 50%

Move the cursor in vertical direction using the keys and, using the numeric key- pad, set the start/finish ramp temperature values (T min and T max) and the required rate of change for the regulator. Pressing the ent key causes the working screen to be activated. Use the key to start ramp generation, starting at the actual dry well temperature and finishing at the T max value. Use the key to start a negative ramp function (actual dry well temperature to T min). The display shows the ramp direction, together with a corresponding heating/cooling symbol.
50.00 ° C
To stop the ramp, press the key. Press the keys to restart. The status bar shows the ramp symbol, the temperature set-point, the symbols and a HOLD message (if the ramp has been stopped).

PRESET
DIRECT
# 1 = 0.00 °C # 2 = 0.00 °C # 3 = 0.00 °C # 4 = 0.00 °C # 5 = 0.00 °C
STEPS RAMP
The next display allows a specific temperature value to be assigned to each of the numeric keys (1 to 5) of the key-pad. Use the keys and the numeric key-pad to set the values. Press ent to leave the set-up display and to open the working screen.
Pressing the # key followed by one of the numeric keys 1- 5 recalls the set-point assigned to the numeric key.
Ramp, Step and Preset functions can also be set starting from the Menu Selection Screen. In an active standard working screen, press the menu key to open the Menu Selection Screen and use the keys to select the required temperature reference from the REF. T display (see next section), and press ent. The following screen allows set-up of the Ramp, Step and Preset parameters.

The DBC series has a dual display. By selecting a measurement (input) option from the menu using the keys and pressing ent, the working screen shows a dual display with the dry well temperature (at the bottom) and the selected input (at the top). Note:
REF T.
T/ C K 23.8° C
RDG: °C 0.15% SPAN 148.50
INTERNAL SEL RTD
NONE
150.00 DIRECT
T/C
BLK °C 148.00
REF T. MEASURE
MEASURE
REF T.
NONE
INTERNAL SEL RTD
NONE
RTD mA · T/C RTD mA

Selecting the Direct, Step or Preset generation mode, the final working screen shows the dry well temperature at the bottom (together with the temperature set- point, the generation mode and the status arrow symbol) and the contact state at the top (open/close). An audio signal indicates the change of state.
150.00 DIRECT
BLK °C 148.00
To enable the Switch Test with the indication of open, close and calculated hysteresis values, the Ramp generation mode must first be selected.
The switch test function can be rapidly recalled from any working screen by pressing the # and +/- keys. Press these keys again to disable Switch Test and to recall only the dry well temperature measurement.
REF T.
NONE
INTERNAL SEL RTD
T/C
RTD mA

After the SET MODE selection, a new display allows the T/C type definition.
K J S N R
Using the keys, select the type of probe to be measured and press ent: the next display sets the compensation method for the T/C cold junction.
COLD J.
This sets the CJ compensation temperature reference to a programmable value between 0 to 70° C. The T/C reference table uses a CJ temperature of 0° C.
CJC °C: 24.2 °C
Using the keys, select the required CJ compensation method and press ent. The display shows the mode settings (i.e. how the T/C measurement can be displayed).

DIRECT CAL
RDG: °C 148.50
150.00 DIRECT
BLK °C 148.00
Using the select the T min value and set the temperature limit using the numeric key-pad: then, move the cursor on the T max value using the key, set the related temperature value and press ent to confirm. Press ent again to show the next display, where the error calculation method can be set.

ABSOLUTE % OF SPAN % OF RDG
Using the keys, select the required error calculation method and press ent.
RDG: °C 0.15% RDG 148.50
150.00 DIRECT
BLK °C 148.00
The working screen shows, in the top part, the selected T/C type, the probe measure directly in the selected temperature unit, the CJ temperature and the error value. The bottom part of the screen shows, the generated temperature value (with internal or external reference), the temperature set-point, the status arrows symbols and the selected generation mode.

RDG: °C 0.15% RDG 148.50
150.00 DIRECT
BLK °C 148.00
The screen shows, in the top part, the selected T/C type, the probe measure directly in the selected temperature unit, the CJ temperature and the error value. The bottom part of the screen shows, the generated temperature value (with internal or external reference), the temperature set-point, the status arrows symbols and the selected generation mode.

REF T.
INTERNAL SEL RTD
NONE
T/C
From the Menu Selection Screen, using the select RTD and press ent. After selection of the generation mode, a new display shows the RTD types that can be measured.
Pt-100 Pt-200 Pt-500 Pt-1000 Ni-100
Using the keys to select the required RTD type and press ent. After this operation, a new display sets the connection configuration (3 or 4 wires) of the probe under test.
3 WIRES 4 WIRES
Using the keys, select the required connection configuration and press ent.
mA RTD

The display shows the mode settings (i.e. how the RTD measurement can be displayed).
DIRECT CAL
150.00 DIRECT
BLK °C 148.00
Using the select the T min value and set the temperature limit using the numeric key-pad: then, move the cursor on the T max value using the key, set the related temperature value and press ent to confirm. Press ent again to show the next display, where the error calculation method can be set.
Pt-100

ABSOLUTE % OF SPAN % OF RDG
Using the keys, select the required method and press ent.
150.00 DIRECT
BLK °C 148.00
The screen shows, in the top part, the selected RTD type, the probe measure directly in the selected temperature unit and the error value. The bottom part of the screen shows, the generated temperature value (with internal or external reference), the temperature set-point, the status arrows symbols and the selected generation mode.
Pt-100

150.00 DIRECT
BLK °C 148.00
The screen shows, in the top part, the selected RTD type, the probe measure directly in the selected temperature unit and the error value. The bottom part of the screen shows, the generated temperature value (with internal or external reference), the temperature set-point, the status arrows symbols and the selected generation mode.
Pt-100

The DBC-TC series has an input section for direct measurement of mA signal, for example, from a DIN temperature transmitter, directly fitted on the probe under test.
REF T.
NONE
INTERNAL SEL RTD
RTD T/C
From the Menu Selection Screen, select mA using the and press ent. After selection of the required generation mode (DIRECT, PRESET, STEP, RAMP), the measure MODE display will be immediately recalled, allowing solution of how the mA measurement must be carried out.
DIRECT CAL
BLK °C 148.00
mA · mA

T min : 50.00 °C T max : 200.00 °C Min : 4. 000 mA Max : 20.000 mA
Using the keys to move the cursor through this display and, using the numeric key-pad, set a low temperature limit (T min) corresponding to a low input current limit (Min) and a high temperature limit (T max) corresponding to the high input current limit (Max) and press ent to confirm the selection and to go to the next display where the error calculation method can be set, as previously described. Selecting ABSOLUTE, the final working screen shows, in the top part, the mA signal from the unit under test, scaled into the selected temperature unit with the error indication reported as deviation in temperature. Selecting % OF SPAN or % OF RDG the display shows the error as a percentage value of full-scale or as a percentage value of specific reading point.
150.00 DIRECT
BLK 148.00 °C
CAL mA

The DBC-TC series can measure a voltage signal from temperature transmitters under test.
REF T.
INTERNAL SEL RTD
From the Menu Selection Screen, select V in the MEASURE section using the keys and press ent. After selection of the temperature generation mode to be used, the next display allows set-up of the voltage measurement method.
DIRECT CAL
RDG: V 10.00
150.00 DIRECT
BLK °C 148.00
mV Ohm V

T min : 50.00 °C T max : 200.00 °C Min : 0 V Max : 10.000 V
Using the keys, move the cursor through this display and set the values with the numeric key-pad: T min value is the low limit for the UUT temperature range, related to its low limit output voltage, T max is the full-scale for the UUT temperature range, related to the high limit of its output voltage. Press ent to confirm and to go to the next display, where the error calculation mode can be set.
ABSOLUTE % OF SPAN % OF RDG
Using the keys, select ABSOLUTE and press ent: the final working screen shows the generated temperature parameters in the bottom part of the display, and the voltage measurement scaled into the selected temperature unit in the top part. The error (deviation) calculation is shown with small digit between the two sections of the display and with a temperature unit value. If in the ERROR screen, using the keys, the % OF SPAN or % OF RDG functions are selected, pressing the ent key, the final working screen shows the error calculation as a percentage deviation related to the full-scale of the device under test temperature range or to its specific measuring point.
RDG: 0,15% SPAN 148.50 °C
150.00 DIRECT
BLK 148.00 °C

The DBC-TC series can measure the low level signal from a thermocouple device under test: using this capability, the operator can make a direct verification of the electrical behaviour for the T/C under test.
REF T.
INTERNAL SEL RTD
From the Menu Selection Screen, select mV using the keys and press ent. After the usual selection of the temperature generation mode to be used (Direct, Ramp, Preset or Step), the final working screen will be immediately recalled showing the generated temperature parameters in the bottom part of the display and the measured millivolt value in the top part of it.
BLK °C 148.00
mV Ohm V · mV

The DBC-TC series can directly measure the resistance value of an RTD under test.
REF T.
INTERNAL SEL RTD
From the Menu Selection Screen, select Ohm in the MEASURE section using the
keys and press ent. After selection of the temperature generation mode, the next display sets the connected RTD as a 3 or 4 wire device.
3 WIRES 4 WIRES
Using the select the required connection and press ent. The final working screen shows the generated temperature in the lower part of the display and the measured Ohmic value in the top part.
Ω
150.00 DIRECT
BLK °C 100.00
mV · V Ohm · Ohm

Using the standard digital RS232 interface, the calibrator can be controlled from a remote computer. By sending commands via the RS232 interface, the calibrator automatically retrieves temperature readings or modifies any parameter settings (temperature set-point, temperature reference selection, ramp/step/preset generation etc.). 4.1 DBC Series - Serial Commands Protocol
aa: Two digit DBC address in ASCII format. (i.e.. 01). Send 00 to overwrite every other address.
par: Parameter. Command depending format.
R: Read Only command
n.a.: Not Applicable
Flow Command Parameters Description Model (TS/TC)
TC TC TC Write Read Query
Par = 0 Absolute error °C 1% of span 2% of readings
R/W - Error Computation Mode (MM command to set CAL mode)
n.a. *aaER? !aaER=par
Par = floating point R – Get Error Value (Mode selected by EM command)
c =1 Regulator (optional) 2 Measure par = 0 NONE
1 n.a. 2 T/C +CJ 3 T/C no CJ 4 RTD 3 Wires 5 RTD 4 Wires 6 RES 3 Wires 7 RES 4 Wires 8 mV 9 n.a. 10 n.a. 11 n.a. 12 mA 13 n.a. 14 Switch 15 n.a. 16 n.a. 17 Reference RTD 4 wires 18 V
Thermocouple + Cold Junction Thermocouple without Cold Junct. Thermoresistors 3 wires Thermoresistors 4 wires Resistors up to 400 Ω 3 wires Resistors up to 400 Ω 4 wires mV input
TC TC TC
TC
TS/TC
TS/TC TS/TC
User Thermoresistors 4 wires Volt Input

Flow Command Parameters Description Model (TS/ TC)
n.a. *aaIRc?
c = 1 Reference RTD 2 Measure par = floating point value
R – Get Input Measure value TS/TC
c = 1 Reference RTD 2 Measure par = 0 NONE
TC TS/TC TS/TC
10 PT100 IEC 11 PT200 IEC 12 PT500 IEC 13 PT1000 IEC 16 Ni100 DIN 19 Reference RTD 20 T/C B IEC 21 T/C E IEC 22 T/C J IEC 23 T/C K IEC 24 T/C N IEC 25 T/C R IEC 26 T/C S IEC 27 T/C T IEC 28 T/C U DIN 29 T/C L DIN 30 T/C C
TC Write Read Query
par = Floating Point R/W –Manual Cold Junction Value (Value in current Temperature Units)
*aaIUc=par *aaIUc ? !aaIUc=par
c = 1 Reference RTD 2 Measure par =98 NONE
TC TS/ TC TS/ TC TS/ TC TS/ TC
Unit n.a Fahrenheit Celsius Kelvin Milliampere Volts Millivolts Ohms
80 °F 81 °C 82 °K 60 mA 61 V 62 mV 63 Ω
TC Write Read Query
par = 0 silent keyboard 1 beep on
R/W –Keyboard bid TS/ TC

Flow Command Parameters Description Model (TS/TC)
*aaKM[ 1]= par *aaKM[ 1]? !aaKM[ 1]= par
par = 0 keys disabled 1 keys enabled
R/ W – Local lockout keyboard TS/TC
*aaMRc= par *aaMRc? !aaMRc= par
c = 1 zero scale 2 full-scale par = floating point value
R/ W – Measure Range (mA or V in CAL ERROR mode)
*aaMM= par *aaMM? !aaMM= par
par = 0 Direct mode 1 Cal mode
R/ W – Measurement mode TC TC
R/ W – Output mode TS/TC Write Read Query
*aaOM[ 1]= par *aaOM[ 1]? !aaOM[ 1]= par
par = 0 Direct mode 1 Preset mode 2 Steps mode 3 Ramp mode
*aaOP[ 1]= par *aaOP[ 1]? !aaOP[ 1]= par
par = floating point value R/ W – Point Temperature TS/TC
n. a. *aaOR[ 1]? !aaOR[ 1]= par
R – Read Controlled Temperature TS/TC
*aaPB[ 1]= par *aaPB[ 1]? !aaPB[ 1]= par
par = 0 to 99.9 floating point value
R/ W – Proportional band TS/TC
*aaPSc= par *aaPSc? !aaPSc= par
par = floating point value c = 1 preset step #1 2 preset step #2 3 preset step #3 4 preset step #4 5 preset step #5
R/ W – Preset Set Values TS/TC TS TC
c = 1 Model 2 Software Version
R – Read Model and Software Version.
*aaRM[ 1]= par *aaRM[ 1]? !aaRM[ 1]= par
par = 0 Ramp Down 1 Ramp Up 2 Down Hold 3 Up Hold
R/ W – Ramp mode TS/TC

Flow Command Parameters Description Model (TS/TC)
*aaRRc= par *aaRRc? !aaRRc= par
par = floating point c = 1 zero scale 2 full-scale
R/W – Ramp Range TS/TC
*aaRS[ 1]= par *aaRS[ 1]? !aaRS[ 1]= par
par r = floating point R/W – Ramp Slope TS/TC
*aaSA[ 1]= par *aaSA[ 1]? !aaSA[ 1]= par
par = 1 to 99 R/W – Station Address TS/TC
*aaSL[ 1]= par *aaSL[ 1]? !aaSL[ 1]= par par = 0 ENGLISH 1 FRENCH 2 GERMAN 3 ITALIAN 4 PORTUGUESE 5 SPANISH
R/W – Language TS/TC
n.a. *aaSN[ 1]? !aaSN[ 1]= par par = 1 to 65535
R – Serial Number TS/TC
*aaSP[ 1]= par *aaSP[ 1]? !aaSP[ 1]= par
par = 1 to 5 - # of step R/W – Number of preset step TS/TC
*aaSWc= par *aaSWc? !aaSWc= par
par = floating point c = 1 – Opened 2 – Closed 3 - Hysteresis
R – Switch test threshold TS/TC
*aaTD[ 1]= par *aaTD[ 1]? !aaTD[ 1]= par
par = 1 to 255 R/W – Derivative constant TS/TC
*aaTI[ 1]= par *aaTI[ 1]? !aaTI[ 1]= par
par = 1 to 255 R/W – Integration constant TS/TC
*aaTU[ 1]= par *aaTU[ 1]? !aaTU[ 1]= par
par = 80 °F 81 °C 82 °K
R/W – Temperature units TS/TC
*aaTRc= par *aaTRc? !aaTRc= par par = floating point value c = 1 low scale 2 full-scale
R/ W – Temperature Range (CAL ERROR mode)

Flow Command Parameters Description Model (TS/TC)
n. a. *aa. Ac? !aa. Ac= par
par = floating point c = 1 regulator channel 2 n. a. 3 calibrator board
R – Get cold junctions compensation TS/TC
n. a. *aa. Mc? !aa. Mc= par
par = Exp. floating point c = 1 regulator temp. 2 REF Probe temp. 3 calibrator value
R – Not averaged measures (diagnostic purposes)
n. a. *aa. Pc? !aa. Pc= par
par = Exp. floating point c = 1 reg. Sensor mV out 2 REF sensor mV out 3 Cal. Electrical value.
R – Main measures
*aaRN = par *aaRN? !aaRN = par
par = integer 0 = REF 1 1 = REF 2
Reference RTD selection TS/TC
n.a. *aaST? !aaST = par
par = integer 0 = Cooling 1 = Heating 2 = Ready
Status of controller TS/TC
*aaCN = par *aaCN? !aaCN = par
par = integer 0 = 10% 1 = 30% 2 = 50% 3 = 70% 4 = 90% 5 = 100%
Select display contrast TS/TC
*aaURc = par *aaURc? !aaURc = par
par = floating c = 1 REF 1 c = 2 REF 2
Ref probe R0 coefficient TS/TC
*aaUAc = par *aaUAc? !aaUAc = par
par = floating c = 1 REF 1 c = 2 REF 2
Ref probe A coefficient TS/TC
*aaUBc = par *aaUBc? !aaUBc = par
par = floating c = 1 REF 1 c = 2 REF 2
Ref probe B coefficient TS/TC
*aaUCc = par *aaUCc? !aaUCc = par
par = floating c = 1 REF 1 c = 2 REF 2
Ref probe C coefficient TS/TC
*aa. Tc *aa. Tc? !aa. Tc= par
par = integer c = 1 REF 1 c = 2 REF 2
Ref probe tag number TS/TC

The DBC series of temperature calibrators must be considered as reference standard: these instruments must be periodically tested and calibrated with a recommended recalibration period of 12 months. The Service Centres of GE provide a global calibration service to National and International standards.
Calibration must be carried out by the suitably qualified calibration technicians. The calibration procedure requires a high-accuracy reference standard (at least 5 times better than the DBC accuracy) with direct traceability to National or International Standard (SIT, UKAS etc.). Press the # and the menu keys to select the set-up display.
CALIBRATION ENTER CODE
ENTER CODE
Use the keys to select CALIBRATION and press ent.
Enter the correct PIN at the ENTER CODE prompt and press ent. After entering the PIN number, calibrate the instrument by following the on-screen instructions.
REF. RTD INPUT
REF. RTD BLOCK
CALIBRATION mV (1) mV (2)

Using a decade resistance box, generate the required value (to the Ref. Probe connection on the electrical panel) and press ent. The microprocessor combines the two values automatically. The nominal value is the value recommended for the test.
Press ENT or CE ...
Tag:
Move the cursor with the keys and set the value with the numeric key-pad. In this way, a characterisation of the specific Reference Probe can be made. Press ent to return to the calibration menu.
accuracy.

1 To calibrate the dry well regulation, install a suitable reference probe in the dry well and connect it to the Reference Probe socket in the front panel.
2 Follow the above procedure to insert the coefficients of the probe in use.
3 Change from INTERNAL to SEL RTD, Direct mode. The display shows the temperature measured by the Reference Probe.
4 Select the CALIBRATION menu and use the PIN to enter the menu. Record the values of SET1 ...SET4, and REF1 ... REF4.
5 Set the dry well temperature at SET1 value; wait at least 30 minutes at this temperature. Record the true temperature value on display, identify this value as READ1.
6 Set the dry well temperature at SET2 value; wait at least 30 minutes at this temperature. Record the true temperature value on display, identify this value as READ2. Repeat this procedure for SET3 and SET4 and record values READ3 and READ4.
7 Enter the Calibration menu, select dry well CALIBRATION.
8 Calculate the NEWREF1 ... NEWREF4 value as follow: NEWREFx= READx- SETx+REFx.
9 Type in REF1 ...REF4 fields the values NEWREF1 ...NEWREF4.
10 Press ENTER. DBC650 (150) automatically calculates the coefficients.
11 The calibration of the dry well temperature regulator has been completed.
REF. RTD INPUT
REF. RTD BLOCK
CALIBRATION mV (1) mV (2)
From the CALIBRATION menu select one of the mV input range to be calibrated using the and press ent.

Using a certified and high accuracy millivolt simulator, generate a millivolt signal to the T/C electrical connection on the front panel (nominal value is the recommended value for the calibration): pressing the ent key causes the microprocessor to combine the millivolt values (true value and effective reading).
Using a certified and high accuracy Volt simulator, apply the calibration signal to the Voltage electrical connection on the front panel of the instrument (the nominal value is the recommended value for the test). Pressing the ent key causes the microprocessor to combine the voltages indication (True value and Reading).
Using a traceable and high accuracy mA simulator, generate a specific current value to the related electrical connection on the front panel of the instrument (the
mV (1) · mA INPUT

Nominal Value is the recommended value for the test): pressing the ent key causes the microprocessor to combine the current indications on the display (True Value and Effective Reading).
Ohms INPUT 1 KOhms INPUT
From the CALIBRATION menu select the required Ohms input range to be calibrated, using the keys, and press ent.
Nominal value: 300 True value: 300.00 Reading: 298.00 Press ENT or CE ...
Ω
Ω
Using a certified and high accuracy Ohms simulator, apply the required value to the related electrical connection on the front panel of the instrument (the nominal value is the recommended value for the test). Pressing the ent key causes the microprocessor to combine the resistance measure on the display (true value and effective reading).
Ohms INPUT

Fault description Possible problem Solution
- faulty protection fuse - faulty power cable - faulty power switch
-Replace the fuse -Replace the power cable -Contact your service representative
- short circuit on power supply card - faulty power switch
-Contact your service representative
- faulty power supply card - faulty controller card - damage to one or more Peltier cells
-Contact your service representative
- faulty controller card -Contact your service representative
-damage to one or more Peltier cells or damage to the heating system
-Contact your service representative
-Contact your service representative
7 At high temperatures the instrument cannot stabilize.
-incorrect step or preset values -faulty reference probe
-Change the step or preset values -Contact your service representative
Fuses are designed to give maximum protection in case of fault. Make sure that voltage selector and fuses are selected for the nominal working value.

1 Wait until the instrument is at ambient temperature before cleaning.
2 Disconnect the instrument from the power supply and clean the keyboard and case with a damp cloth, using a little mild detergent. CAUTION:
DO NOT, under any circumstances, use any form of solvent for cleaning purposes. Using solvent damages the case.

DBC 150-TC and TS DBC 650-TC and TS
Temperature range -45°C (from ambient) to +150°C 50°C to 650°C Well depth dry well 160mm (6.3in) insertion depth 155mm (6.1in)
dry well 160mm (6.3in) insertion depth 155mm (6.1in) Well diameter 30mm (1.18in) 30mm (1.18in) Stability 0.03°C 0.05°C (between 70°and 650°C) 0.15°C (between 50°and 70°C) Radial Uniformity (adjacent holes)
Heating time 14 minutes (20°C to 120°C) 22 minutes (25°C to 650°C) Cooling time 22 minutes (25°C to -20°C) 27 minutes (600°C to 100°C with optional cooling probe) Display resolution 0.01°C 0.01°C CJ compensation error 0.3°C 0.3°C Reference probe accuracy 0.2°C (including B1probe) 0.3°C (including B2 probe) Power supply 100-120 or 220-240 Vac, 50-60 Hz, 230 VA 100-120 or 220-240 Vac, 50-60 Hz, 1385 VA DBC 150 TC only DBC 650 TC only Maximum mA IN (current signal) 35 mA 35 mA Maximum V IN (voltage signal) 30 Vdc 30 Vdc
Display area 60 x 40mm (2.36 x 1.57 in) LCD with backlighting Temperature units °C K °F External Ref. Probe Facility for connection of a PT100 reference probe Switch Test Facility for connection of switch contacts Communication Interface RS232, supplied as standard Software function Ramp, step and preset temperature generation. Switch test Operating temperature range 10 to 50°C Reference Standards
EN 61326 (1997) + A1(1998) + A2 (2001) EN 61010-1: 2001 (2nd edition) Measurement Category I (all measurement terminals MAX 30V to earth) Pollution degree 2 Regulation sensor Thermocouple Weight 9.5 kg (21 lbs) Dimensions 322 x 156 x 328mm (12.7 x 6 x 12.9 in)
{DBC-TC only} Thermocouples · EMC Safety

Input Range Resolution Accuracy
PT100 -99 /+750°C 0.01°C 0.04°C 0.15°C PT200 -99 /+750°C 0.1°C 0.1°C 0.3°C PT500 -99 /+750°C 0.1°C 0.1°C 0.3°C PT1000 -99 /+550°C 0.1°C 0.1°C 0.3°C Ni100 -60 /+250°C 0.01°C 0.05°C 0.15°C
T/C K -99 /+990°C 0.1°C 0.1°C 0.3°C T/C J -99 /+990°C 0.1°C 0.1°C 0.3°C T/C S -50 /+100°C 0.1°C 0.3°C 1°C 100 /+990°C 0.1°C 0.2°C 0.6°C T/C N -99 /+990°C 0.1°C 0.1°C 0.3°C T/C R -50 /+100°C 0.1°C 0.3°C 0.9°C 100 /+990°C 0.1°C 0.2°C 0.6°C T/C B 200 /+500°C 0.1°C 0.8°C 2.5°C 500 /+990°C 0.1°C 0.5°C 1.5°C T/C C 0 /+990°C 0.1°C 0.2°C 0.6°C T/C E -99 /+990°C 0.1°C 0.1°C 0.3°C T/C T -99 /+400°C 0.1°C 0.1°C 0.3°C T/C L -99 /+800°C 0.1°C 0.1°C 0.3°C 800 /+900°C 0.1°C 0.3°C 0.9°C T/C U -99 /+50°C 0.1°C 0.2°C 0.6°C -50 /+0°C 0.1°C 0.1°C 0.3°C 0 /+500°C 0.1°C 0.1°C 0.3°C 500 /+600°C 0.1°C 0.3°C 0.9°C
Volts 0 /+12V 0.001V 0.01%RDG+0.01%FRO 0.03%RDG+0.03%FRO mV 0 /+79.2mV 0.001mV 0.003%RDG+0.002%FRO 0.01%RDG+0.006%FRO mA 0 /+24mA 0.001mA 0.005%RDG+0.005%FRO 0.015%RDG+0.015%FRO Ohm 0 /+400ߟ 0.01ߟ 0.003%RDG+0.003%FRO 0.01%RDG+0.01%FRO
Accuracy · (90 days) · (1 year)



Common problems, likely causes, and what to do.
| Problem | Possible cause | Remedy | |
|---|---|---|---|
| The instrument does not work even if the power cord is correctly fitted and the mains switch is in the correct position. | -Replace the fuse -Replace the power cable -Contact your service representative | p.63 | |
| When the instrument is turned on the fuse blows | -Contact your service representative | p.63 | |
| The heating and cooling function of the dry well is not working | -Contact your service representative | p.63 | |
| The controller is not disabled when the set- point is reached | -Contact your service representative | p.63 | |
| The heating and cooling times out of specification. | -Contact your service representative | p.63 | |
| Fan system not working. | -Contact your service representative | p.63 | |
| At high temperatures the instrument cannot stabilize. | -Change the step or preset values -Contact your service representative | p.63 |