A&D GP-61K

98 pages·2.7 MB·Building Set·PDF·ENDEFRITNLJA
A&D GP-61K
Table of Contents 80 sections WM+PD4000273D · 1. INTRODUCTION · 1-1 About This Manual · 1-2 Features · … ShowHide
WM+PD4000273Dp.11. INTRODUCTIONp.51-1 About This Manualp.51-2 Featuresp.51-3 Compliancep.7Compliance with FCC Rulesp.7Compliance with EMC Directivesp.72. UNPACKING THE BALANCEp.92-1 Unpackingp.9Standard type: GP-12K/20K/30K/32K/40K/60K/61K/100K/102Kp.9Separate display type: GP-30KS/32KS/60KS/61KS/100KSp.102-2 Installing the Balancep.103. PRECAUTIONSp.113-1 Before Usep.113-3 After Usep.123-4 Power Supplyp.123-5 Display Symbols and Key Operationp.13Key operationp.13Display symbolsp.133-6 Smart Range Functionp.14Precision range/standard range valuep.144. WEIGHING UNITSp.154-1 Unitsp.154-2 Changing the Unitsp.185. WEIGHINGp.195-1 Basic Operation (Kilogram Mode)p.195-2 Counting Mode (PCS)p.20Selecting the counting modep.20Storing a sample unit mass (Weighing input mode)p.20Counting operationp.21Counting mode using the ACAI functionp.215-3 Percent Mode (%)p.22Selecting the percent modep.22Storing the 100% reference massp.22Reading the percentagep.225-5 Accumulation Functionp.23Selecting the unitp.23Setting the function tablep.23Using the accumulation functionp.24Outputting the valuep.246. RESPONSE ADJUSTMENT / SELF CHECK FUNCTIONp.256-1 Automatic Response Adjustment / Self Check Functionp.25Operationp.256-2 Manual Response Adjustmentp.267. CALIBRATIONp.277-1 Calibration Groupp.27Displayp.287-2 Automatic Self Calibration (Calibration due to changes in temperature)p.287-3 Calibration Using the Internal mass (One-Touch Calibration)p.29About the internal massp.297-4 Calibration Using an External Weightp.307-5 Calibration Test Using an External Weightp.327-6 Correcting the Internal Mass Valuep.34Example: Using the GP-20Kp.348. FUNCTION SWITCH AND INITIALIZATIONp.368-1 Permit or Inhibitp.36Switchesp.368-2 Initializing the Balancep.379. FUNCTION TABLEp.389-1 Structure and Sequence of the Function Tablep.38Examplep.389-2 Display and Keysp.389-3 Details of the Function Tablep.399-4 Description of the Class “Environment, Display”p.42Condition ( Cond )p.429-5 Description of the Item “Data output mode”p.44Key modep.44Auto print modes A and Bp.44Stream modep.44Interval memory modep.459-6 Description of the Item “Data format”p.459-7 Description of the Data Format Added to the Weighing Datap.47Tare valuep.47Comparison resultsp.479-8 Data Format Examplesp.489-9 Clock and Calendar Functionp.509-10 Comparator Functionp.52Setting example 1p.52Setting example 2p.539-11 Adding the Comparison Resultsp.54
Page 1 of the A&D GP-61K user manual — WM+PD4000273D
Page 1 · click to enlarge
Page 1
A&D GP-61K — WM+PD4000273D diagram (page 1)
Fig. 1
A&D GP-61K — WM+PD4000273D diagram (page 1)
Fig. 2

WM+PD4000273D

Page 2 of the A&D GP-61K user manual — WM+PD4000273D
Page 2 · click to enlarge
Page 2

© 2006 A&D Company Ltd. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into any language in any form by any means without the written permission of A&D Company Ltd. The contents of this manual and the specifications of the instrument covered by this manual are subject to change for improvement without notice.

Page 3 of the A&D GP-61K user manual — WM+PD4000273D
Page 3 · click to enlarge
Page 3
Page 3 of the A&D GP-61K user manual — WM+PD4000273D
Page 4 of the A&D GP-61K user manual — WM+PD4000273D
Page 4 · click to enlarge
Page 4
Page 4 of the A&D GP-61K user manual — WM+PD4000273D
Page 5 of the A&D GP-61K user manual — 1-2 Features
Page 5 · click to enlarge
Page 5

1. INTRODUCTION

This manual describes how the GP series balance works and how to get the most out of it in terms of performance.

Read this manual thoroughly before using the balance and keep it at hand for future reference.

1-1 About This Manual

This manual consists of the following five parts:

Basic operation Describes precautions on handling the balance, balance

construction and basic balance operation.

Adapting to the environment Describes response adjustment, calibration and calibration

test.

Functions Describes various functions of the balance.

RS-232C serial interface Describes RS-232C serial interface output and external contact

input. The RS-232C serial interface outputs weighing data and controls the balance. The external contact input commands the balance re-zeroing and data output.

Maintenance Describes maintenance, error codes, troubleshooting, specifications and options.

1-2 Features

z Large Vacuum Fluorescent Display, easy to read.

z Dust-tight and Protected Against Water Jets (Complying with IP65), allows washing with water.

z Built-in Calibration Weight (hereinafter referred to as internal mass), allows easy calibration,

adjustment and maintenance of the balance.

z Automatic Self Calibration, using the internal mass, adapting to changes in temperature.

z Self Check Function, provided to self-check the balance using the internal mass.

z Automatic Response Adjustment, adapting to vibration and drafts in the environment.

z High Response Speed, the time to read a displayed value after a sample is placed on the pan

has been shortened by using the Super Hybrid Sensor (SHS). (Approx. 1.5 seconds when FAST is selected for the response rate.)

z Data Memory Function, storing weighing data, calibration data or unit mass in the counting

mode. It can also store tare values or upper and lower limit values for the comparator function. Interval Memory Mode is provided to store the weighing data periodically.

z Good Laboratory Practice (GLP) data output using the standard RS-232C serial interface.

z Clock and Calendar Function, adding the time and date to the output data.

z Comparator Indicators, displaying the comparison results, H I, OK or LO.

z Capacity Indicator, displaying the weight value in percentage relative to the weighing capacity.

Page 6 of the A&D GP-61K user manual — 1. INTRODUCTION
Page 6 · click to enlarge
Page 6

z Hold Function, provided for weighing a moving object such as an animal.

z Multiple Weighing Units, with most of the common units used around the world.

z Density Mode, for calculating the density of a solid.

z Accumulation Function, adding the weight values and outputting the sum.

z Adjustable Display Position, the standard type balance has a swing-arm to adjust the display

position for optimum viewing. Separate display type balances (GP-30KS/32KS/60KS/61KS/100KS) are also available.

z Windows Communication Tools (WinCT), allows easy communication with Windows.

Windows is a registered trademark of Microsoft Corporation.

z Reference Card, provided for a quick reference to the balance operation.

z Underhook, available as an option, for measuring density and weighing magnetic materials.

z Comparator Output (OP-04) and Analog Voltage Output (OP-06) are available as options.

Page 7 of the A&D GP-61K user manual — Compliance with EMC Directives
Page 7 · click to enlarge
Page 7

1-3 Compliance

Compliance with FCC Rules

Please note that this equipment generates, uses and can radiate radio frequency energy. This equipment has been tested and has been found to comply with the limits of a Class A computing device pursuant to Subpart J of Part 15 of FCC rules. These rules are designed to provide reasonable protection against interference when equipment is operated in a commercial environment. If this unit is operated in a residential area, it may cause some interference and under these circumstances the user would be required to take, at his own expense, whatever measures are necessary to eliminate the interference.

(FCC = Federal Communications Commission in the U.S.A.)

Compliance with EMC Directives

This device features radio interference suppression in compliance with valid

EC Regulation 89/336/EEC.

Page 8 of the A&D GP-61K user manual — 1-3 Compliance
Page 8 · click to enlarge
Page 8
A&D GP-61K — 1-3 Compliance diagram (page 8)
Fig. 3 · 1-3 Compliance
A&D GP-61K — 1-3 Compliance diagram (page 8)
Fig. 4 · 1-3 Compliance
A&D GP-61K — 1-3 Compliance diagram (page 8)
Fig. 5 · 1-3 Compliance
Page 9 of the A&D GP-61K user manual — Standard type: GP-12K/20K/30K/32K/40K/60K/61K/100K/102K
Page 9 · click to enlarge
Page 9

2. UNPACKING THE BALANCE

2-1 Unpacking

z The balance is a precision instrument. Unpack the balance carefully. Keep the packing

material to be used for transporting the balance in the future.

z The packing contents depend on the balance model. See the illustrations to confirm that

everything is contained.

Standard type: GP-12K/20K/30K/32K/40K/60K/61K/100K/102K

A&D GP-61K — Standard type: GP-12K/20K/30K/32K/40K/60K/61K/100K/102K diagram (page 9)
Fig. 6 · Standard type: GP-12K/20K/30K/32K/40K/60K/61K/100K/1…

Please confirm that the AC adapter type is correct for your local voltage and receptacle type.

Page 10 of the A&D GP-61K user manual — 2-2 Installing the Balance
Page 10 · click to enlarge
Page 10

Separate display type: GP-30KS/32KS/60KS/61KS/100KS

A&D GP-61K — Separate display type: GP-30KS/32KS/60KS/61KS/100KS diagram (page 10)
Fig. 7 · Separate display type: GP-30KS/32KS/60KS/61KS/100KS

Please confirm that the AC adapter type is correct for your local voltage and receptacle type.

2-2 Installing the Balance

Install the balance as follows:

1. Refer to “3. PRECAUTIONS” for installing the balance.
2. Adjust the leveling feet to level the balance. Confirm it using the bubble spirit level.
3. Confirm that the adapter type is correct for the local voltage and power receptacle type.
4. Connect the AC adapter to the balance. Warm up the balance for at least 30 minutes with nothing on the weighing pan.
Page 11 of the A&D GP-61K user manual — 3-1 Before Use
Page 11 · click to enlarge
Page 11

3. PRECAUTIONS

To get the optimum performance from the balance and acquire accurate weighing data, note the

following:

3-1 Before Use

z Install the balance in an environment where the

temperature and humidity are not excessive. The best operating temperature is about 20°C / 68°F at about 50% relative humidity.

z Install the balance where it is not exposed to direct sunlight

and it is not affected by heaters or air conditioners.

z Install the balance where it is free of dust.

A&D GP-61K — 3-1 Before Use diagram (page 11)
Fig. 8 · 3-1 Before Use

z Install the balance away from equipment which

produces magnetic fields.

z Install the balance in a stable place avoiding vibration

and shock. Corners of rooms on the first floor are best, as they are less prone to vibration.

z The weighing table should be solid and free from

vibration, drafts and as level as possible.

z Level the balance by adjusting the leveling feet and

confirm it using the bubble spirit level.

z Ensure a stable power source when using the AC adapter.

z Plug in the AC adapter and warm up the balance for at

least 30 minutes.

z Calibrate the balance periodically for accurate weighing.

z When the balance is installed for the first time or has been

moved, warm up the balance for at least 12 hours to allow the balance to reach equilibrium with the ambient temperature, and then perform calibration before use.

A&D GP-61K — 3-1 Before Use diagram (page 11)
Fig. 9 · 3-1 Before Use

z Discharge static electricity from the material to be

weighed (hereinafter referred to as sample). When a sample could have a static charge, the weighing data is influenced. Try to keep the ambient humidity above 45%RH or use the metal shield case.

Page 12 of the A&D GP-61K user manual — 3-4 Power Supply
Page 12 · click to enlarge
Page 12

z This balance uses a strong magnet as part of the

A&D GP-61K — 3-3 After Use diagram (page 12)
Fig. 10 · 3-3 After Use

balance assembly, so please use caution when weighing magnetic materials such as iron. If there is a problem, use the underhook on the bottom of the balance to suspend the material away from the influence of the magnet.

z Cancel the temperature difference between a sample

and the environment. When a sample is warmer (cooler) than the ambient temperature, the sample will be lighter (heavier) than the true weight. This error is due to a rising (falling) draft around the sample.

A&D GP-61K — 3-3 After Use diagram (page 12)
Fig. 11 · 3-3 After Use

z Make each weighing gently and quickly to avoid errors

due to changes in the environmental conditions.

z Do not drop things upon the weighing pan, or place a

sample on the pan that is beyond the balance weighing capacity. Place a sample in the center of the weighing pan.

z Do not use a sharp instrument such as a pencil to

press the keys. Use your finger only.

z Press the RE-ZERO key before each weighing to

prevent possible errors.

z Calibrate the balance periodically so as to cancel

possible errors.

z Take into consideration the affect of air buoyancy on a

sample when more accuracy is required.

z Keep the balance interior free of dust and foreign

materials.

3-3 After Use

z Avoid mechanical shock to the balance.

z Do not disassemble the balance. Contact the local A&D dealer if the balance needs service or repair.

z Do not use organic solvents to clean the balance. Clean the balance with a lint free cloth that

is moistened with warm water and a mild detergent.

z Do not allow the balance to be immersed in water. Even though the balance complies with

IP65 (Dust-tight and Protected Against Water Jets), the balance will not withstand being completely immersed in water.

3-4 Power Supply

z Do not remove the AC adapter while the internal mass is in motion, for example, right after the

AC adapter is connected, or during calibration using the internal mass. If the AC adapter is removed under the conditions described above, the internal mass will be left unsecured, that may cause mechanical damage when the balance is moved. Before removing the AC adapter, press the ON:OFF key and confirm that zero is displayed.

z When the AC adapter is connected, the balance is in the standby mode if the standby indicator

is on (refer to “3-5 Display Symbols and Key Operation”). This is a normal state and does not harm the balance. For accurate weighing, keep the AC adapter connected to the balance and AC power unless the balance is not to be used for a long period of time.

Page 13 of the A&D GP-61K user manual — Display symbols
Page 13 · click to enlarge
Page 13

3-5 Display Symbols and Key Operation

Key operation

Key operation affects how the balance functions. The basic key operations are:

z “Press and release the key immediately” or “Press the key”

= normal key operation during measurement

z “Press and hold the key”

Display symbols

A&D GP-61K — Display symbols diagram (page 13)
Fig. 12 · Display symbols

Each key, when pressed or when pressed and held, functions as follows:

Key When pressed When pressed and held

Turns the display ON and OFF. The standby indicator is displayed when the display is turned off. The weighing mode is enabled when the display is turned on. This key is available anytime. Pressing the key during operation will interrupt the operation and turn the display OFF.

In the weighing mode, turns the minimum weighing value ON and OFF.

Enters the function table mode. Refer to “9. FUNCTION TABLE”.

In the counting or percent mode, enters the sample storing mode.

Switches the weighing units stored in the function table. Refer to “4. WEIGHING UNITS”.

Performs response adjustment and self check.

Performs calibration of the balance using the internal mass.

Displays other items of the calibration menu.

Stores the weighing data in memory or outputs to a printer or personal computer depending on the function table settings. (Factory setting = output)

No function at the factory setting

By changing the function table: Outputs “Title block” and “End block” for GLP report. Displays the data memory menu. Sets the display to zero.

Page 14 of the A&D GP-61K user manual — Precision range/standard range value
Page 14 · click to enlarge
Page 14

3-6 Smart Range Function

The GP-32K, GP-32KS and GP-102K are equipped with two ranges. The precision range has a

higher resolution. The standard range has normal resolution.

The range is switched automatically, depending on the value displayed. Pressing the RE-ZERO key

allows weighing in the precision range, regardless of the tare value. (Smart range function)

The range can be fixed to the standard range, by pressing the SAMPLE key.

Below is a description of how the Smart Range actually functions, using the GP-102K

(Precision range=61 kg×1 g, Standard range=101 kg×10 g)

A&D GP-61K — 3-6 Smart Range Function diagram (page 14)
Fig. 13 · 3-6 Smart Range Function

1 Press the RE-ZERO key. The balance will start weighing, using the precision range.

2 Place a container, on the weighing pan, to be used as a tare. The value displayed exceeds the precision range value. So, the range will be switched to the standard range.

3 Press the RE-ZERO key. The balance will be switched to the precision range.

4 Place a sample on the pan. The value displayed is within the precision range value. So, the balance will perform a weighing, using the precision range.

Precision range/standard range value

A&D GP-61K — Precision range/standard range value diagram (page 14)
Fig. 14 · Precision range/standard range value

Standard range 6.101 to 31.008 kg 61.01 to 101.08 kg

Page 15 of the A&D GP-61K user manual — 4-1 Units
Page 15 · click to enlarge
Page 15

4. WEIGHING UNITS

4-1 Units

With the GP series balance, the following weighing units and weighing modes are available :

A&D GP-61K — 4-1 Units diagram (page 15)
Fig. 15 · 4-1 Units

A unit or mode can be selected and stored in the function table as described in “4-2 Changing the Units”. If a weighing mode (or unit of weight) has been turned off, that mode or unit will be missing in the sequence. Tael has four varieties, one of which can be selected and installed at the factory.

For details about the units and modes, see the table below:

A&D GP-61K — 4-1 Units diagram (page 15)
Fig. 16 · 4-1 Units

Name (unit, mode) Abbrev. Display Function table (Storing mode)

1 g = Gram g G 1 g kilogram kg 1000 g Counting mode PCS Percent mode Ounce (Avoir) OZ 28.349523125 g Pound Lb 453.59237 g Pound/Ounce L OZ l0 1Lb=16 oz, 1 oz=28.349523125 g Troy Ounce OZt 31.1034768 g Metric Carat ct 0.2 g Momme mom 3.75 g Pennyweight dwt 1.55517384 g Tael (HK general, Singapore) 37.7994 g Tael (HK jewelry) 37.429 g Tael (Taiwan) 37.5 g Tael (China)

31.25 g Tola (India) t 11.6638038 g Messghal MS 4.6875 g Density mode

Page 16 of the A&D GP-61K user manual — 4. WEIGHING UNITS
Page 16 · click to enlarge
Page 16

The tables below indicate the weighing capacity and the minimum display for each unit, depending

on the balance model.

A&D GP-61K — 4. WEIGHING UNITS diagram (page 16)
Fig. 17

Ounce (Avoir) 423.290 740.755 1093.495 2151.710 0.005 1446.24 0.02

Pound/Ounce 26L 7.29 46L 4.75 68L 5.49 134L 7.7 0.01 90L 6.24 0.02

Troy Ounce 385.810 675.165 996.675 1961.195 0.005 1318.18 0.02

Metric Carat 60000.0 105000.0 155000.0 305000.0 0.5 205000 5

Tael (HK general, Singapore) 317.465 555.565 820.120 1613.780 0.005 1084.68 0.02

Tael (HK jewelry) 320.605 561.060 828.235 1629.750 0.005 1095.40 0.02

Tael (Taiwan) 320.000 560.000 826.665 1626.665 0.005 1093.34 0.02

Tael (China) 384.000 672.000 992.000 1952.000 0.005 1312.00 0.02

Tola (India) 1028.82 1800.44 2657.80 5229.85 0.01 3515.15 0.05

A&D GP-61K — 4. WEIGHING UNITS diagram (page 16)
Fig. 18
Page 17 of the A&D GP-61K user manual — 4. WEIGHING UNITS
Page 17 · click to enlarge
Page 17
A&D GP-61K — 4. WEIGHING UNITS diagram (page 17)
Fig. 19
A&D GP-61K — 4. WEIGHING UNITS diagram (page 17)
Fig. 20
Page 18 of the A&D GP-61K user manual — 4-2 Changing the Units
Page 18 · click to enlarge
Page 18

4-2 Changing the Units

The units or modes can be selected and stored in the function table. The sequence of displaying the units or modes can be arranged so as to fit the frequency of use in the function table.

Select a unit or mode and arrange the sequence of display as follows:

1 Press and hold the SAMPLE key until ba5fnc of the function table is displayed, then release the key.

2 Press the SAMPLE key several times to display Unit .

A&D GP-61K — 4-2 Changing the Units diagram (page 18)
Fig. 21 · 4-2 Changing the Units

3 Press the PRINT key to enter the unit selection mode.

4 Specify a unit or mode in the order to be displayed using the following keys.

SAMPLE key To sequentially display the units.

RE-ZERO key To specify a unit or mode. The stabilization indicator appears when the displayed unit or mode is specified.

5 Press the PRINT key to store the units or modes. The balance displays end and then displays the next menu item of the function table.

6 Press the CAL key to exit the function table. Then the balance returns to the weighing mode with the selected unit.

Page 19 of the A&D GP-61K user manual — 5-1 Basic Operation (Kilogram Mode)
Page 19 · click to enlarge
Page 19

5. WEIGHING

5-1 Basic Operation (Kilogram Mode)

A&D GP-61K — 5-1 Basic Operation (Kilogram Mode) diagram (page 19)
Fig. 22 · 5-1 Basic Operation (Kilogram Mode)

1 Place a container on the weighing pan, if necessary. Press the RE-ZERO key to cancel the weight (tare). The balance displays 0.0000 kg. (The decimal point position depends on the balance model.)

2 Place a sample on the pan or in the container.

3 Wait for the stabilization indicator to be displayed. Read the value.

4 Remove the sample and container from the pan.

Press the SAMPLE key to turn on or off the minimum weighing value.

The weighing data can be stored in memory. For details, refer to “11. DATA MEMORY”.

Page 20 of the A&D GP-61K user manual — Storing a sample unit mass (Weighing input mode)
Page 20 · click to enlarge
Page 20

5-2 Counting Mode (PCS)

This is the mode to determine the number of objects in a sample based on the standard sample unit mass. Unit mass means the mass of one sample. The smaller the variables in each sample unit mass is, the more accurate the counting will be. The GP series balance is equipped with the Automatic Counting Accuracy Improvement (ACAI) function to improve the counting accuracy.

A&D GP-61K — 5-2 Counting Mode (PCS) diagram (page 20)
Fig. 23 · 5-2 Counting Mode (PCS)

Selecting the counting mode

1 Press the MODE key to select pcs (counting mode).

Storing a sample unit mass (Weighing input mode)

2 Press the SAMPLE key to enter the sample unit mass storing mode.

3 To select the number of samples, press the SAMPLE key several times. It may be set to 10, 25, 50 or 100.

A greater number of samples will yield more accurate counting result.

4 Place a container on the weighing pan, if necessary. Press the RE-ZERO key to cancel the weight (tare). The number specified in step 3 appears. e.g.: 25 0 pcs is displayed if 25 is selected in step 3.

5 Place the number of samples specified on the pan. In this example, 25 pieces.

6 Wait for the stabilization indicator to come on. Press the PRINT key to calculate and store the unit mass. The balance displays 25 pcs (counting mode) and is set to count samples with this unit mass. (The sample unit mass stored, even if the AC adapter is removed, is maintained in non-volatile memory.)

To improve the accuracy of the unit mass, proceed to step 8.

If the balance judges that the mass of the samples is too light to aquire accurate weighing, it displays an error requiring the addition of more samples to the specified number. In the example above, 150- pcs appears, requiring 25 more samples. Add 25 samples and press the PRINT key. When the unit mass is stored correctly, the balance proceeds to the counting mode.

Key entry of the unit mass value (digital input mode) is possible. For details, refer to “11-4 Memory for Unit Mass in the Counting Mode”

Page 21 of the A&D GP-61K user manual — Counting mode using the ACAI function
Page 21 · click to enlarge
Page 21

Counting operation

7 Place the samples to be counted on the pan.

Counting mode using the ACAI function

The ACAI is a function that improves the accuracy of the unit mass automatically by increasing the number of samples as the counting process proceeds.

8 If a few more samples are added, the processing indicator turns on. To prevent an error, add three or more. The processing indicator does not turn on if overloaded. Try to add the same number of samples as displayed.

9 The balance re-calculates the unit mass while the processing indicator is blinking. Do not touch the balance or samples on the pan until the processing indicator turns off.

10 Counting accuracy is improved when the processing indicator turns off.

Each time the above operation is performed, a more accurate unit mass will be obtained. There is no definite upper limit of ACAI range for the number of samples exceeding 100. Try to add the same number of samples as displayed.

11 Remove all the samples used in ACAI and proceed with the counting operation using the improved unit mass.

ACAI will not function on the unit mass entered using the keys, or digital input mode.

Page 22 of the A&D GP-61K user manual — Reading the percentage
Page 22 · click to enlarge
Page 22

5-3 Percent Mode (%)

This is the mode to display the weight value in percentage compared to a 100% reference mass and

is used for target weighing or checking the sample variable.

Selecting the percent mode

A&D GP-61K — Selecting the percent mode diagram (page 22)
Fig. 24 · Selecting the percent mode

1 Press the MODE key to select % (percent mode). If the percent mode can not be selected, refer to “4. WEIGHING UNITS”.

Storing the 100% reference mass

2 Press the SAMPLE key to enter the 100% reference mass storing mode. Even in the storing mode, pressing the MODE key will switch to the next mode.

3 Place a container on the weighing pan, if necessary. Press the RE-ZERO key to cancel the weight (tare). The balance displays 100 0 % .

4 Place the sample to be set as the 100% reference mass on the pan or in the container.

5 Press the PRINT key to store the reference mass. The balance displays 100.00 % . (The decimal point position depends on the reference value. The reference mass stored, even if the AC adapter is removed, is maintained in non-volatile memory.)

6 Remove the sample.

Reading the percentage

7 Place a sample to be compared to the reference mass on the pan. The displayed percentage is based on the 100% reference mass. 5-4 Animal Weighing Mode (Hold Function)

This is the mode to weigh a moving object such as an animal, even when the display of the weighing data fluctuates. The hold function allows the average weight of the animal to be displayed. To use the hold function, set the function in the function table. For details, refer to “9. FUNCTION TABLE” and “9-4 Description of the Class “Environment, Display””.

Page 23 of the A&D GP-61K user manual — Setting the function table
Page 23 · click to enlarge
Page 23

5-5 Accumulation Function

The accumulation function adds the weighing data and displays the total value.

To use the accumulation function, set the “Accumulation function (add)” parameter of the function table as described below.

Selecting the unit

1 Press the MODE key to select a unit to be used for accumulation.

While the accumulation function is in use, unit selection using the MODE key is not available.

Setting the function table

2 Press and hold the SAMPLE key until ba5fnc of the function table is displayed, then release the key.

3 Press the PRINT key.

4 Press the SAMPLE key several times to display add 0 .

5 Press the RE-ZERO key to display add 1 .

6 Press the PRINT key to store the setting.

7 Press the CAL key to return to the weighing mode.

Page 24 of the A&D GP-61K user manual — Outputting the value
Page 24 · click to enlarge
Page 24

Using the accumulation function

Use the keys below to operate the accumulation function.

MODE key: Displays the weighing data and the total value alternately each time it is pressed. Will not change the unit while the accumulation function is in use.

RE-ZERO key: Sets the display to zero while the weighing data is displayed. Deletes the total value while the total value is displayed.

PRINT key Outputs and adds the weighing data while the weighing data is displayed. Outputs the total value while the total value is displayed.

A&D GP-61K — Using the accumulation function diagram (page 24)
Fig. 25 · Using the accumulation function

1 Press the RE-ZERO key to set the display to zero.

2 Place a sample on the pan. The weight value is displayed.

3 Press the PRINT key. The weight value is added to the total and output. The accumulation number at the upper left of the display increases by one.

4 Repeat steps 1-3 to accumulate more data.

5 Press the MODE key to display the total value.

Outputting the value

6 Press the PRINT key to output the total value.

7 Press the RE-ZERO key to delete the total value.

A&D GP-61K — Outputting the value diagram (page 24)
Fig. 26 · Outputting the value

While the accumulation function is in use, the data memory function is not available.

To disable the accumulation funciton, set the “Accumulation function (add)” parameter to “0”.

When the “Data number output (d-no)” parameter is set to “1”, the accumulation number will be output before the weighing data.

Page 25 of the A&D GP-61K user manual — Operation
Page 25 · click to enlarge
Page 25

6. RESPONSE ADJUSTMENT / SELF CHECK FUNCTION

This function detects the influence on weighing that is caused by drafts and/or vibration at the place where the balance is installed and sets the response characteristic automatically. When this function is selected, the balance self-checks the performance at the same time.

Two modes of response adjustment are available: automatic and manual.

A&D GP-61K — 6. RESPONSE ADJUSTMENT / SELF CHECK FUNCTION diagram (page 25)
Fig. 27

Changing the response rate changes the display refresh rate.

Indicator Parameter Response characteristic Display refresh rate FAST Cond 0 MID. Cond 1 SLOW Cond 2

If the response rate is changed as follows: MID. or SLOW FAST =10 times/second FAST MID. or SLOW = 5 times/second

6-1 Automatic Response Adjustment / Self Check Function

This function automatically updates the response adjustment by analyzing the influence of the environment on the weighing data and also self-checks the balance performance using the internal mass.

Operation

A&D GP-61K — Operation diagram (page 25)
Fig. 28 · Operation

1 Press and hold the MODE key until RESPONSE is displayed, then release the key.

2 The balance automatically starts to check the balance performance and sets the response characteristic.

3 After automatic adjustment, the balance displays the updated response indicator and returns to the weighing mode. The response indicator remains displayed for a while.

e.g. “ MID. OK ”

The example above indicates that the result of the self check is good and MID. is selected as the response rate.

Page 26 of the A&D GP-61K user manual — Operation
Page 26 · click to enlarge
Page 26

If the automatic response adjustment fails, the balance displays CH ng. Check the ambient conditions such as breeze and vibration, also check the weigning pan. Then, perform the adjustment again. To return to the weighing mode, press the CAL key.

If the automatic response adjustment is awkward, try to refine it using the manual response adjustment.

6-2 Manual Response Adjustment

This function manually updates the response adjustment.

Operation

A&D GP-61K — Operation diagram (page 26)
Fig. 29 · Operation

1 Press and hold the MODE key until RESPONSE is displayed, then release the key.

And then, press the MODE key again quickly.

2 Press the MODE key to select a rate of the response adjustment. Either FAST, MID. or SLOW can be selected.

3 After a few seconds of inactivity the balance displays eend . Then, it returns to the weighing mode and displays the updated response indicator. The response indicator remains displayed for a while.

Page 27 of the A&D GP-61K user manual — 7-1 Calibration Group
Page 27 · click to enlarge
Page 27

7. CALIBRATION

7-1 Calibration Group

The GP series balance has the following modes as a calibration group.

Calibration z Automatic self calibration (calibration due to changes in temperature) z Calibration using the internal mass (one-touch calibration) z Calibration using an external weight Calibration test z Calibration test using an external weight (Calibration test does not

perform calibration.) Correction of the internal mass value Terms

The following terms are defined as follows: Internal mass = Built-in calibration weight External weight = A weight that you have. Referred to as a calibration weight when used for calibration. Calibration weight = A weight used for calibration Target weight = An external weight used for calibration test Caution

  • z Calibration adjusts the balance for accurate weighing. Besides periodic calibration and before each use, perform calibration when:
  • the balance is installed for the first time.
  • the balance has been moved.
  • the ambient environment has changed.

z Do not allow vibration or drafts to affect the balance during calibration.

z To output the data for GLP using the RS-232C interface, set “GLP output (info)” of “Data

output (dout)”. For details, refer to “9. FUNCTION TABLE”. Time and date are added to the GLP report. If the time or date is not correct, adjust them. For details, refer to “9-9 Clock and Calendar Function”.

z Calibration test is available only when “GLP output (info)” of “Data output (dout)” is set to “1” or “2”,

z The calibration and calibration test data can be stored in memory. To store them, set “Data

memory (data)” to “3”. For details, refer to “11. DATA MEMORY”. Caution on using an external weight

z The accuracy of an external weight will influence the accuracy of weighing. Select an

appropriate weight as listed below:

Model Usable calibration weight Adjustable range GP-12K 5 kg, 10 kg GP-20K 10 kg, 20 kg GP-30K/GP-30KS 20 kg, 30 kg GP-32K/ GP-32KS 20 kg, 30 kg GP-40K 20 kg, 30 kg, 40 kg

-1.5 g to +1.5 g

GP-60K/ GP-60KS 30 kg, 40 kg, 50 kg, 60 kg -15 g to +15 g GP-61K/ GP-61KS 30 kg, 40 kg, 50 kg, 60 kg -1.5 g to +1.5 g GP-100K/GP-100KS 60 kg, 80 kg, 100 kg GP-102K 60 kg, 80 kg, 100 kg -15 g to +15 g

The calibration weight in bold type: factory setting The calibration weight value can be adjusted within the range above.

Page 28 of the A&D GP-61K user manual — 7-2 Automatic Self Calibration (Calibration due to changes in temperature)
Page 28 · click to enlarge
Page 28

Display

z This indicator means “the balance is measuring calibration data”.

Do not allow vibration or drafts to affect the balance while this indicator is displayed.

7-2 Automatic Self Calibration (Calibration due to changes in temperature)

This function automatically calibrates the balance when the balance detects an ambient temperature change. If GLP output is selected in the function table, the balance outputs the calibration report or stores the data in memory. Automatic self calibration functions even if the display is turned off (standby state).

The displays shown below are related to the automatic self calibration.

Indicates that the balance detects a change in ambient temperature and automatic self calibration will start. If the balance is not used for a few minutes with this indicator blinking, the balance performs automatic self calibration. The blinking duration depends on the environment.

Indicates that the balance is measuring calibration data. Do not allow vibration or drafts to affect the balance while this indicator is displayed. After calibration, the balance returns to indicate the previous display.

Page 29 of the A&D GP-61K user manual — About the internal mass
Page 29 · click to enlarge
Page 29

7-3 Calibration Using the Internal mass (One-Touch Calibration)

This function calibrates the balance using the internal mass. The only operation required is to press

the CAL key.

Operation

1 Plug in the AC adapter and warm up the balance for at least 30 minutes with nothing on the weighing pan.

2 Press the CAL key.

3 The balance displays Calin and performs calibration using the internal mass. Do not allow vibration or drafts to affect the balance

4 The balance displays end after calibration. If the “GLP output (info)” parameter of the function table is set to “1” or “2”, the balance displays glp and outputs “Calibration Report” using the RS-232C interface or stores the data in memory. For details on the calibration report format, refer to “10-2 GLP Report”.

5 The balance will automatically return to the weighing mode after calibration.

About the internal mass

The mass of the internal mass may change due to corrosion or other damage caused by the operating environment, or due to aging. Check the internal mass periodically. Correct the internal mass value as necessary. For details, refer to “7-6 Correcting the internal mass value”.

To maintain the weighing accuracy, perform the calibration using an external weight periodically, as described below.

Page 30 of the A&D GP-61K user manual — Operation
Page 30 · click to enlarge
Page 30

7-4 Calibration Using an External Weight

This function calibrates the balance using an external weight.

A&D GP-61K — 7-4 Calibration Using an External Weight diagram (page 30)
Fig. 30 · 7-4 Calibration Using an External Weight

Operation

1 Plug in the AC adapter and warm up the balance for at least 30 minutes with nothing on the pan.

2 Press and hold the CAL key until Calout is displayed, then release the key.

3 The balance displays Cal 0 .

z If you want to change the calibration weight (a

list of usable weights is shown on page 25), press the SAMPLE key and proceed to step 4.

z If you use the calibration weight value stored in

the balance, proceed to step 5.

4 Specify the calibration weight value as follows:

SAMPLE key To switch the display condition to: “All of the segments blinking” (calibration weight selection mode) or “The last two digits blinking” (value adjustment mode).

RE-ZERO key To select the calibration weight or adjust the value. In the value adjustment mode, -15 digits appear after +15 digits.

PRINT key To store the new weight

value. Even if the AC adapter is removed, the data is maintained in non-volatile memory.

CAL key To cancel the operation and

return to Cal 0 .

Page 31 of the A&D GP-61K user manual — 7-4 Calibration Using an External Weight
Page 31 · click to enlarge
Page 31
A&D GP-61K — 7-4 Calibration Using an External Weight diagram (page 31)
Fig. 31 · 7-4 Calibration Using an External Weight

5 Confirm that there is nothing on the pan and press the PRINT key. The balance measures the zero point. Do not allow vibration or drafts to affect the balance.

The balance displays the calibration weight value.

6 Place the displayed calibration weight on the pan and press the PRINT key. The balance measures the calibration weight. Do not allow vibration or drafts to affect the balance.

7 The balance displays end . Remove the weight from the pan.

8 If the “GLP output (info)” parameter, of the function table, is set to “1” or “2”, the balance displays glp and outputs “Calibration Report” using the RS-232C interface or stores the data in memory. For details on the calibration report format, refer to “10-2 GLP Report”.

9 The balance will automatically return to the weighing mode.

10 Place the calibration weight on the pan and confirm that the value displayed is within ±2 digits of the specified value. If it is not within the range, check the ambient conditions such as breeze and vibration, also check the weighing pan. Then, repeat steps 1 to 10.

Page 32 of the A&D GP-61K user manual — Operation
Page 32 · click to enlarge
Page 32

7-5 Calibration Test Using an External Weight

This function tests the balance weighing accuracy using an external mass and outputs the result. This is available only when the “GLP output (info)” parameter is set to “1” or “2”. (Calibration test does not perform calibration.)

A&D GP-61K — 7-5 Calibration Test Using an External Weight diagram (page 32)
Fig. 32 · 7-5 Calibration Test Using an External Weight

Operation

1 Connect the AC adapter and warm up the balance for at least 30 minutes with nothing on the pan.

2 Press and hold the CAL key until CCout is displayed, then release the key.

3 The balance displays CC 0 . z If you want to change the target weight (a list of

usable weights is shown on page 25), press the SAMPLE key and proceed to step 4. z If you use the target weight value stored in the

balance, proceed to step 5.

4 Specify the target weight value as follows: SAMPLE key To switch the display condition to: “All of the segments blinking” (target weight selection mode) or “The last two digits blinking” (value adjustment mode).

RE-ZERO key To select the target weight or

adjust the value. In the value adjustment mode, -15 digits appear after +15 digits.

PRINT key To store the new weight

value. Even if the AC adapter is removed, the data is maintained in non-volatile memory.

CAL key To cancel the operation and

return to CC 0 .

Page 33 of the A&D GP-61K user manual — 7-5 Calibration Test Using an External Weight
Page 33 · click to enlarge
Page 33
A&D GP-61K — 7-5 Calibration Test Using an External Weight diagram (page 33)
Fig. 33 · 7-5 Calibration Test Using an External Weight

5 Confirm that there is nothing on the pan and press the PRINT key. The balance measures the zero point and displays the measured value. Do not allow vibration or drafts to affect the balance. The balance displays the target weight value.

6 Place the displayed target weight on the pan and press the PRINT key. The balance measures the target weight and displays the measured value. Do not allow vibration or drafts to affect the balance.

7 The balance displays end . Remove the weight from the pan.

8 The balance displays glp and outputs “Calibration Test Report” using the RS-232C interface or stores the calibration test data in memory. For details on the calibration test report format, refer to “10-2 GLP Report”.

9 The balance will automatically return to the weighing mode.

Page 34 of the A&D GP-61K user manual — Example: Using the GP-20K
Page 34 · click to enlarge
Page 34

7-6 Correcting the Internal Mass Value

The GP series balance can correct the internal mass value within the range shown below. This function corrects the internal mass value to conform to an external weight. The corrected mass value is maintained in non-volatile memory even if the AC adapter is removed. The internal mass value is corrected as follows:

A&D GP-61K — 7-6 Correcting the Internal Mass Value diagram (page 34)
Fig. 34 · 7-6 Correcting the Internal Mass Value

Model Correction reference value Correction range GP-12K 10 kg GP-20K GP-30K/GP-30KS GP-32K/ GP-32KS GP-61K/ GP-61KS

GP-60K/ GP-60KS 60 kg -50 g to +50 g GP-100K/GP-100KS

Example: Using the GP-20K

A&D GP-61K — Example: Using the GP-20K diagram (page 34)
Fig. 35 · Example: Using the GP-20K
Page 35 of the A&D GP-61K user manual — Operation
Page 35 · click to enlarge
Page 35

Operation

1 Calibrate the balance using the internal mass. (one-touch calibration) Then, place an external weight and confirm the value to be corrected.

In the example, the value is to be corrected by 0.6 gram in 20 kilograms.

2 Press the ON:OFF key to turn off the display.

3 While pressing and holding the PRINT key and the SAMPLE key, press the ON:OFF key. The balance displays p5 .

4 Press the PRINT key. Then the balance displays the function switches.

Set the function table switch and internal mass correction switch to “1” as shown above using the following keys.

SAMPLE key To select the switch to change the

value. The selected digit blinks. RE-ZERO key To change the parameter of the

switch selected.

5 Press the PRINT key to store the new setting. The balance returns to the weighing mode.

6 Press and hold the SAMPLE key to enter the function table and release the key when ba5fnc is displayed.

7 Press the SAMPLE key several times until C5in is displayed, then release the key.

8 Press the PRINT key to enter the procedure for correcting the internal mass value.

9 Correct the internal mass value using the following keys.

RE-ZERO key To select the value. (-50 digits appear after +50 digits.) PRINT key To store the new value and display the

next menu item of the function table. CAL key To cancel the correction and

display the next menu item of the function table. 10 Press the CAL key. The balance returns to the weighing mode.

11 Press the CAL key to calibrate the balance using the internal mass.

12 Place the external weight on the pan and confirm that the correction has been performed properly. In this example, confirm that the value displayed is within ±2 digits of the correction reference value or 20 kilograms. If the value is not within ±2 digits of the correction value, repeat the above procedure to correct it.

Page 36 of the A&D GP-61K user manual — Operation
Page 36 · click to enlarge
Page 36

8. FUNCTION SWITCH AND INITIALIZATION

8-1 Permit or Inhibit

The balance stores parameters that must not be changed unintentionally (e.g. Calibration data for accurate weighing, Data for adapting to the operating environment, Control data for the RS-232C interface). There are five switches for the purpose of protecting these parameters. Each switch can select either “permit” or “inhibit”. The “inhibit” protects parameters against unintentional operations.

Switches

A&D GP-61K — Switches diagram (page 36)
Fig. 36 · Switches

Operation

1 Press the ON:OFF key to turn off the display.

2 While pressing and holding the PRINT key and the SAMPLE key, press the ON:OFF key. The balance displays p5 .

3 Press the PRINT key. Then the balance displays the function switches.

4 Set the switches using the following keys.

SAMPLE key To select the switch to change the parameter. The selected switch blinks.

RE-ZERO key To change the parameter of the switch selected.

0:To inhibit changes./ Can not be used.

1:To permit changes./ Can be used.

PRINT key To store the new parameter and return to the weighing mode.

CAL key To cancel the operation and return to the weighing mode.

Page 37 of the A&D GP-61K user manual — Operation
Page 37 · click to enlarge
Page 37

8-2 Initializing the Balance

This function returns the following parameters to factory settings.

z The sample unit mass value (counting mode), 100% reference mass value (percent mode)

z The data that is stored in the balance using the data memory function

z External calibration weight and target weight value

z Function switch settings

z Liquid density and temperature in the density mode

A&D GP-61K — 8-2 Initializing the Balance diagram (page 37)
Fig. 37 · 8-2 Initializing the Balance

Be sure to calibrate the balance after initialization.

Operation

1 Press the ON:OFF key to turn off the display.

2 While pressing and holding the PRINT key and the SAMPLE key, press the ON:OFF key. The balance displays p5 .

3 Press the SAMPLE key to display Clr .

4 Press the PRINT key. To cancel this operation, press the CAL key.

5 Press the RE-ZERO key.

6 Press the PRINT key to initialize the balance.

The balance will automatically return to the weighing mode.

z Calibration data · z Function table

Page 38 of the A&D GP-61K user manual — 9-2 Display and Keys
Page 38 · click to enlarge
Page 38

9. FUNCTION TABLE

The function table reads or rewrites the parameters that are stored in the balance. These parameters are maintained in non-volatile memory, even if the AC adapter is removed.

9-1 Structure and Sequence of the Function Table

The function table menu consists of two layers. The first layer is the “Class” and the second layer is the “Item”. Each item stores a parameter.

Example

This example sets “Stores weighing data” for “Data memory” and “1 minute” for “Interval time”.

A&D GP-61K — Example diagram (page 38)
Fig. 38 · Example

9-2 Display and Keys

Display/Key Description The symbol “ ” indicates that the parameter displayed is in effect. When pressed and held in the weighing mode, enters the function table mode. Selects the class or item in the function table mode. Changes the parameter. When a class is displayed, moves to an item in the class. When an item is displayed, stores the new parameter and displays the next class. When an item is displayed, cancels the new parameter and displays the next class. When a class is displayed, exits the function table mode and returns to the weighing mode.

Page 39 of the A&D GP-61K user manual — 9-3 Details of the Function Table
Page 39 · click to enlarge
Page 39

9-3 Details of the Function Table

A&D GP-61K — 9-3 Details of the Function Table diagram (page 39)
Fig. 39 · 9-3 Details of the Function Table
Page 40 of the A&D GP-61K user manual — 9-3 Details of the Function Table
Page 40 · click to enlarge
Page 40
A&D GP-61K — 9-3 Details of the Function Table diagram (page 40)
Fig. 40 · 9-3 Details of the Function Table
Page 41 of the A&D GP-61K user manual — 9-3 Details of the Function Table
Page 41 · click to enlarge
Page 41
A&D GP-61K — 9-3 Details of the Function Table diagram (page 41)
Fig. 41 · 9-3 Details of the Function Table
Page 42 of the A&D GP-61K user manual — Condition ( Cond )
Page 42 · click to enlarge
Page 42

9-4 Description of the Class “Environment, Display”

Condition ( Cond )

Cond 0 This parameter is for sensitive response to the fluctuation of a weight value.

Used for powder target weighing, weighing a very light sample or when quick response weighing is required. After setting, the balance displays FAST.

Cond 2 This parameter is for stable weighing with slow response. Used to prevent a weight value from drifting due to vibration or drafts. After setting, the balance displays SLOW.

With “Hold function (Hold)” set to “ON (1)”, this item is used to set the averaging time.

Stability band width ( 5t-b )

This item controls the width to regard a weight value as a stable value. When the fluctuation per second is less than the parameter, the balance displays the stabilization indicator and outputs or stores the data. The parameter influences the “Auto print mode”

5t-b 0 This parameter is for sensitive response of the stabilization indicator. Used

for exact weighing. 5t-b 2 This parameter ignores slight fluctuation of a weight value. Used to prevent

a weight value from drifting due to vibration or drafts.

Hold function ( Hold ) (Animal weighing mode)

This function is used to weigh a moving object such as an animal. When the weighing data is over the weighing range from zero and the display fluctuation is within the stabilization range for a fixed period of averaging time, the processing indicator illuminates and the balance displays the average weight of the animal. When the animal is removed from the weighing pan, the display returns to zero automatically. This function is available only when the hold function parameter is set to “1” (the animal mode indicator ANIMAL illuminates) and any weighing unit other than the counting mode is selected.

The stabilization range and averaging time are set in “Condition (Cond)” and “Stability band width (5t-b)”.

Weighing range Averaging time Stabilization range GP-12K/20K Cond 0 5t-b 0 Cond 1 5t-b 1 GP-30K/30KS/32K/32KS/ 40K/61K/61KS 20 g Cond 2

2 seconds Faster 4 seconds | 8 seconds More accurate 5t-b 2

| Big GP-60K/60KS/100K/100KS GP-102K 50 g

Small

Page 43 of the A&D GP-61K user manual — 9-4 Description of the Class “Environment, Display”
Page 43 · click to enlarge
Page 43

Zero tracking ( trc )

This function tracks zero point drift caused by changes in the environment and stabilizes the zero point. When the weighing data is only a few digits, turn the function off for accurate weighing.

trc 0 The tracking function is not used. Used for weighing a very light sample. trc 1 The tracking function is used. Normal zero tracking. trc 2 The tracking function is used. Strong zero tracking. trc 3 The tracking function is used. Very strong zero tracking.

Display refresh rate ( 5pd )

Period to refresh the display. This parameter influences “Baud rate”, “Data output pause” and “Stream mode”.

Decimal point ( pnt )

The decimal point format can be selected.

Auto display-ON ( p-on )

When the AC adapter is plugged in, the display is automatically turned on without the ON:OFF key operation, to display the weighing mode. Used when the balance is built into an automated system. 30-minute warm up is necessary for accurate weighing.

Auto display-OFF ( poff )

When the AC adapter is connected and no operation is performed (inactivity state) for 10 minutes, the display is automatically turned off and the standby indicator illuminates.

Capacity indicator ( g5i )

In the weighing mode, the indicator displays the weighing data relative to the weighing capacity in percentage. (Zero = 0%, maximum capacity = 100%) When the “Data memory (data)” parameter is set to “1” (to store unit mass in the counting mode), “2” (to store the weighing data), “4” (to store comparator settings) or “5” (to store tare value), the indicator displays the information stored in memory, such as the amount of memory data or data number.

Accumulation function ( add )

The accumulation function adds the weighing data, displays and outputs the total value. For details, refer to “5-5 Accumulation Function.”

Display at start ( rng )

When the weighing accuracy is not so strict, the smallest displayable weighing value can be turned off without any key operation at weighing start. Useful when the balance is built into an automated system.

Page 44 of the A&D GP-61K user manual — Stream mode
Page 44 · click to enlarge
Page 44

9-5 Description of the Item “Data output mode”

The parameter setting of “Data output mode (prt)” applies to the performance when the “Data memory (data)” parameter is set to “2” (to store the weighing data) and when the data is transmitted using the RS-232C interface.

Key mode

When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs or stores the weighing data and the display blinks one time.

Required setting dout prt 0 Key mode

Auto print modes A and B

When the displayed value is stable and the conditions of “Auto print polarity”, “Auto print difference” and reference value are met, the balance outputs or stores the weighing data. When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs or stores the data and the display blinks one time.

Mode A: Required setting dout prt 1 Auto print mode A (reference = zero) dout ap-p Auto print polarity dout ap-b Auto print difference

Example “For weighing each time a sample is placed and removed, with “ar-d” set to “1” (to adjust zero after the data is output).”

Mode B: Required setting dout prt 2 Auto print mode B (reference = last stable value) dout ap-p Auto print polarity dout ap-b Auto print difference

Example “For weighing while a sample is added.”

Stream mode

The balance outputs the weighing data continuously regardless of the display condition. The display does not blink in this mode. This mode is not available and the interval memory mode is used when the “Data memory (data)” parameter is set to “2” (to store the weighing data).

Required setting dout prt 3 Stream mode dout data 0 Data memory function is not used. ba5fnc 5pd Display refresh rate 5if bp5 Baud rate

Example “For monitoring data on a computer”

Page 45 of the A&D GP-61K user manual — 9-6 Description of the Item “Data format”
Page 45 · click to enlarge
Page 45

Interval memory mode

The weighing data is periodically stored in memory.

Required setting dout prt 3 Interval memory mode dout data 2 Data memory function is used. Stores weighing data. dout int Interval time Optional setting dout 5-td1, 2, or 3 Adds the time and date.

Example “For periodical weighing without a computer command and to output all of the data, to a computer, at one time”

9-6 Description of the Item “Data format”

A&D standard format 5if type 0

This format is used when the peripheral equipment can receive the A&D format. If an AD-8121B is used, set the printer to MODE 1 or 2.

z This format consists of fifteen characters excluding the terminator.

z A header of two characters indicates the balance condition.

z The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus

sign is used.

z The unit, consisting of three characters, follows the data.

A&D GP-61K — 9-6 Description of the Item “Data format” diagram (page 45)
Fig. 42 · 9-6 Description of the Item “Data format”

DP (Dump print) format 5if type 1

This format is used when the peripheral equipment can not receive the A&D format. If an AD-8121B is used, set the printer to MODE 3.

z This format consists of sixteen characters excluding the terminator.

z A header of two characters indicates the balance condition. No overload header is used.

z The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is

not zero or overloaded.

z The unit, consisting of three characters, follows the data.

A&D GP-61K — 9-6 Description of the Item “Data format” diagram (page 45)
Fig. 43 · 9-6 Description of the Item “Data format”
Page 46 of the A&D GP-61K user manual — Interval memory mode
Page 46 · click to enlarge
Page 46

KF format 5if type 2

This is the Karl-Fischer moisture meter format and is used when the peripheral equipment can only communicate using this format.

z This format consists of fourteen characters excluding the terminator.

z This format has no header characters.

z The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is

not zero or overloaded.

z This format outputs the unit only for a stable value.

A&D GP-61K — Interval memory mode diagram (page 46)
Fig. 44 · Interval memory mode

MT format 5if type 3

z A header of two characters indicates the balance condition.

z The polarity sign is used only for negative data.

z The weighing data uses spaces in place of the leading zeros.

z The character length of this format changes dependent upon the unit

A&D GP-61K — Interval memory mode diagram (page 46)
Fig. 45 · Interval memory mode

NU (numerical) format 5if type 4

This format outputs only numerical data.

z This format consists of nine characters excluding the terminator.

z The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus

sign is used.

CSV format 5if type 5

z Separates the data of A&D standard format and the unit by a comma ( , ).

z Outputs the unit even when the data is overloaded.

z When ID number, data number, time and date are added, outputs ID number, data number,

date, time and weighing data in this order and separates each item by a comma and treats all the items as one group of data.

A&D GP-61K — Interval memory mode diagram (page 46)
Fig. 46 · Interval memory mode
Page 47 of the A&D GP-61K user manual — Comparison results
Page 47 · click to enlarge
Page 47

9-7 Description of the Data Format Added to the Weighing Data

ID number dout 5-id 1

The number to identify a specific balance.

z This format consists of seven characters excluding the terminator.

Data number dout d-no 1

This format outputs the data number just before the data is transmitted using the RS-232C interface.

z This format consists of six characters excluding the terminator.

z When CSV format (5if type 5) is selected, the period ( . ) is replaced with a comma ( , ).

Date dout 5-td 2 or 3

z The date output order can be changed in “Clock (Cl adj)”. Outputs the year in four-digit

format.

Time dout 5-td 1 or 3

z Outputs time in 24-hour format.

Tare value

z When the tare value in memory is recalled, the tare value is output before the weighing data.

Weighing data (Net value)

Comparison results

z By setting “Comparison results (Cp-r)” of the function table to “1”, the comparison results can be

added to the data output using the RS-232C serial interface. Use A&D standard format (type 0).

The comparison results are added after the header in A&D standard format as below.

A&D GP-61K — Comparison results diagram (page 47)
Fig. 47 · Comparison results

When the data described above is added to the weighing data, the output is in the following order: ID number, Data number, Date, Time and Weighing data.

Tare value

Page 48 of the A&D GP-61K user manual — 9-8 Data Format Examples
Page 48 · click to enlarge
Page 48

9-8 Data Format Examples

A&D GP-61K — 9-8 Data Format Examples diagram (page 48)
Fig. 48 · 9-8 Data Format Examples
A&D GP-61K — 9-8 Data Format Examples diagram (page 48)
Fig. 49 · 9-8 Data Format Examples
A&D GP-61K — 9-8 Data Format Examples diagram (page 48)
Fig. 50 · 9-8 Data Format Examples
A&D GP-61K — 9-8 Data Format Examples diagram (page 48)
Fig. 51 · 9-8 Data Format Examples
Page 49 of the A&D GP-61K user manual — 9-8 Data Format Examples
Page 49 · click to enlarge
Page 49
A&D GP-61K — 9-8 Data Format Examples diagram (page 49)
Fig. 52 · 9-8 Data Format Examples
A&D GP-61K — 9-8 Data Format Examples diagram (page 49)
Fig. 53 · 9-8 Data Format Examples
A&D GP-61K — 9-8 Data Format Examples diagram (page 49)
Fig. 54 · 9-8 Data Format Examples

When “Pound Ounce” is selected, the data is output with the unit of ounce (oz).

Page 50 of the A&D GP-61K user manual — Operation
Page 50 · click to enlarge
Page 50

9-9 Clock and Calendar Function

The balance is equipped with a clock and calendar function. When the “GLP output (info)” parameter is set to “1” or “2” and the “Time/Date output (5-td)” parameter is set to “1”, “2” or “3”, the time and date are added to the output data. Set or confirm the time and date as follows:

Operation

1 Press and hold the SAMPLE key until ba5fnc of the function table is displayed, then release the key.

2 Press the SAMPLE key several times to display Cl adj .

3 Press the PRINT key. The balance enters the mode to confirm or set the time and date.

4 The current time is displayed with all the digits blinking. z When the time is correct and the date does not

A&D GP-61K — Operation diagram (page 50)
Fig. 55 · Operation

need to be confirmed, press the CAL key and proceed to step 8. z When the time is correct and the date is to be

confirmed, press the SAMPLE key and proceed to step 6. z When the time is not correct and is to be changed,

press the RE-ZERO key and proceed to step 5.

Setting the time (with part of the digits blinking)

5 Set the time in 24-hour format using the following keys.

SAMPLE key To select the digits to change

the value. The selected digits blink. RE-ZERO key To increase the value by one. MODE key To decrease the value by one. PRINT key To store the new setting, display end and go to step 6. CAL key To cancel the new setting and

go to step 6.

Confirming the time

Page 51 of the A&D GP-61K user manual — 9-9 Clock and Calendar Function
Page 51 · click to enlarge
Page 51

6 The current date is displayed with all the digits blinking. z To change the display order of year (y), month (n)

and day (d), press the MODE key. The date is output in the order as specified. z When the date is correct and the operation is to be

finished, press the CAL key and proceed to step 8. z When the time is to be confirmed again, press the

SAMPLE key and go back to step 4. z When the date is not correct and is to be changed,

press the RE-ZERO key and proceed to step 7.

The year is expressed using a two-digit format. For example, the year 2000 is expressed as “00”.

A&D GP-61K — 9-9 Clock and Calendar Function diagram (page 51)
Fig. 56 · 9-9 Clock and Calendar Function

Setting the date (with part of the digits blinking)

7 Set the date using the following keys.

SAMPLE key To select the digits to change

the value. The selected digits blink.

RE-ZERO key To increase the value by one.

MODE key To decrease the value by one.

PRINT key To store the new setting, display end and go to step 8.

CAL key To cancel the new setting and

go to step 8.

8 The balance displays the next menu item of the function table. Press the CAL key to exit the clock and calendar function and return to the weighing mode.

Do not enter invalid values such as a non-existing date when setting the time and date.

When the clock backup battery has been depleted, the balance displays rtc pf . Under this condition, press any key and set the time and date. The dead battery only affects the clock and calendar function. Even so, the function works normally as long as the AC adapter is connected to the balance.

Confirming the date · Quitting the operation

Page 52 of the A&D GP-61K user manual — Setting example 1
Page 52 · click to enlarge
Page 52

9-10 Comparator Function

The results of the comparison are indicated by HI OK LO on the display.

Operating conditionsz No comparison

z Comparison when the weighing data is stable or overloaded, excluding “near zero”

z Comparison when the weighing data is stable or overloaded, including “near zero”

z Continuous comparison, excluding “near zero”

z Continuous comparison, including “near zero”

To compare, usez Upper limit value and lower limit value

z Reference value and tolerance value

Input methodz Digital input

z Weighing input For a description of “Comparator mode (Cp fnc)”, refer to “9-3 Details of the Function Table”. “Near zero” means that the weighing data is within ±10 digits from zero point. Example: In case of GP-30K, the range of ±1.0 g is “near zero”.

Setting example 1

(Comparison when the weighing data is stable or overloaded, excluding “near zero”, upper limit and lower limit)

Selecting a comparator mode

1 Press and hold the SAMPLE key until ba5fnc of the function table is displayed, then release the key.

2 Press the SAMPLE key several times to display Cp fnc .

3 Press the PRINT key.

4 Press the RE-ZERO key several times to display Cp 1 .

5 Press the SAMPLE key to display Cp in .

6 Press the RE-ZERO key several times to display Cp in 0 .

7 Press the PRINT key to store the selected mode.

Entering the upper and lower limit values

8 With Cp Hi displayed, press the PRINT key. The current setting of the upper limit value is displayed with all of the digits blinking.

z When the current setting is not to be changed, press the PRINT or CAL key to proceed to step 9.

z When the current setting is to be changed, press the RE-ZERO key. The balance is now in

the digital input mode. To use the weighing input mode, press and hold the MODE key.

Digital input mode Change the setting using the following keys.

SAMPLE key To select the digit to change the value. RE-ZERO key To change the value of the digit selected. MODE key To switch the polarity. PRINT key To store the new setting and go to step 9. CAL key To cancel the new setting and go to step 9.

Page 53 of the A&D GP-61K user manual — Setting example 2
Page 53 · click to enlarge
Page 53

Press the RE-ZERO key. The balance displays 0.0 g. Place a sample, with a mass that

corresponds to the upper limit value, on the pan. Press the PRINT key to store the upper limit

value. Remove the sample. The balance displays Cp lo .

9 With Cp lo displayed, press the PRINT key. The current setting of the lower limit value is

displayed with all of the digits blinking.

z When the current setting is not to be changed, press the PRINT or CAL key to proceed to step 10.

z When the current setting is to be changed, press the RE-ZERO key. The balance is now in

the digital input mode. To use the weighing input mode, press and hold the MODE key.

Enter the lower limit value in the same way as described in step 8. Then, go to step 10.

10 Press the CAL key to exit the comparator function and return to the weighing mode.

Setting example 2

(Continuous comparison, including “near zero”, reference value and tolerance value) Selecting a comparator mode

1 Press and hold the SAMPLE key until ba5fnc of the function table is displayed, then release the key.

2 Press the SAMPLE key several times to display Cp fnc .

3 Press the PRINT key.

4 Press the RE-ZERO key several times to display Cp 4 .

5 Press the SAMPLE key to display Cp in .

6 Press the RE-ZERO key several times to display Cp in 1 .

7 Press the PRINT key to store the selected mode.

Entering the reference and tolerance values

8 With Cp ref displayed, press the PRINT key. The current setting of the reference value is displayed with all the digits blinking.

z When the current setting is not to be changed, press the PRINT or CAL key to proceed to step 9.

z When the current setting is to be changed, press the RE-ZERO key. The balance is now in

the digital input mode. To use the weighing input mode, press and hold the MODE key.

Digital input mode Change the setting using the following keys.

SAMPLE key To select the digit to change the value. RE-ZERO key To change the value of the digit selected. MODE key To switch the polarity. PRINT key To store the new setting and go to step 9. CAL key To cancel the new setting and go to step 9.

Weighing input mode Press the RE-ZERO key. The balance displays 0.0 g. Place a sample, with a mass that corresponds to the reference value, on the pan. Press the PRINT key to store the reference value. Remove the sample and go to step 9.

Weighing input mode

Page 54 of the A&D GP-61K user manual — Selecting a unit
Page 54 · click to enlarge
Page 54

9 With Cp lnt displayed, press the PRINT key. The current setting of the tolerance value is displayed with all the digits blinking.

z When the current setting is not to be changed, press the PRINT or CAL key to proceed to step10.

z When the current setting is to be changed, press the RE-ZERO key. The balance is now in

the digital input mode. Change the setting using the following keys.

SAMPLE key To select the digit to change the value. RE-ZERO key To change the value of the digit selected. PRINT key To store the new setting and go to step 10. CAL key To cancel the new setting and go to step 10.

10 Press the CAL key to exit the comparator function and return to the weighing mode. Notes When Pound/Ounce is selected as a weighing unit, enter the values in ounces for comparison.

In the density mode, comparison is performed to the density obtained.

9-11 Adding the Comparison Results

By setting “Comparison results (Cp-r)” of the function table to “1”, the comparison results can be

added to the data output using the RS-232C serial interface. Use A&D standard format (type 0). The comparison results are added after the header in A&D standard format as below.

A&D GP-61K — 9-11 Adding the Comparison Results diagram (page 54)
Fig. 57 · 9-11 Adding the Comparison Results

9-12 Main Display Comparison Function

The main display comparison function displays the comparison results in a magnified way, on the

main portion of the display in place of the weight value.

Selecting a unit

1 Press the MODE key to select a unit to be used for comparison.

Page 55 of the A&D GP-61K user manual — Using the main display comparison function
Page 55 · click to enlarge
Page 55

Setting the function table

2 Press and hold the SAMPLE key until ba5fnc of the function table is displayed, then release the key.

3 Press the SAMPLE key several times to display Cp fnc .

4 Press the PRINT key.

5 Press the SAMPLE key several times to display Cp-b 0 .

6 Press the RE-ZERO key to display Cp-b 1 .

To disable the main display comparison funciton, set the “Main display comparison (Cp-b)” parameter to “0”.

7 Press the PRINT key to store the setting.

8 Press the CAL key to return to the weighing mode.

Setting the comparator values

Setting the comparator values as described in the previous section. Setting example (Cp 3=Continuous comparison, excluding “near zero”)

A&D GP-61K — Setting the comparator values diagram (page 55)
Fig. 58 · Setting the comparator values

Using the main display comparison function

1 Press the RE-ZERO key to set the display to zero.

2 Place a sample on the pan. The balance performs a comparison using the specified comparison values and displays the comparison results, HI, OK or LO.

3 Each time the MODE key is pressed, the balance switches between the standard display and the main display comparison. Note that “g0” appears for OK.

If the comparison is not performed, for example, because the weight value is near zero or unstable, the balance displays the weight value even when the main display comparison function is used.

Even while the main display comparison function is in use, the balance re-zeroing and data output is possible.

Only the unit selected before this function can be used. While the main display comparison function is in use, the data memory function is not available.

Page 56 of the A&D GP-61K user manual — Display character set
Page 56 · click to enlarge
Page 56

10. ID NUMBER AND GLP REPORT

z The ID number is used to identify the balance when Good Laboratory Practice (GLP) is used.

z The ID number is maintained in non-volatile memory even if the AC adapter is removed.

z The GLP output format is selected at “GLP output (info)” of the function table and can be

output to a personal computer or printer using the RS-232C serial interface.

z The GLP output format includes the balance manufacturer, model, serial number, ID number,

date, time and space for signature for weighing data, the weight used and results for calibration or calibration test data.

z The balance can output the following reports for GLP.

“Calibration report” of the calibration, using the internal mass (Calibration due to changes in temperature and one-touch calibration.)

“Calibration report” of the calibration, using an external weight.

“Calibration test report” of the calibration test, using an external weight.

“Title block” and “End block” for the weighing data.

z Calibration and calibration test data can be stored in memory to output several reports at the

same time. Refer to “11. DATA MEMORY” for details.

z For details on confirming and setting the time and date, refer to “9-9 Clock and Calendar

Function”.

10-1 Setting the ID Number

1 Press and hold the SAMPLE key until ba5fnc of the function table is displayed, then release the key.

2 Press the SAMPLE key several times to display id .

3 Press the PRINT key. Set the ID number using the following keys.

RE-ZERO key To set the character of the digit selected. Refer to the display character

set shown below.

SAMPLE key To select the digit to change the value.

PRINT key To store the new ID number and display ba5fnc.

CAL key To cancel the new ID number and display ba5fnc.

4 With ba5fnc displayed, press the CAL key to return to the weighing mode.

Display character set

A&D GP-61K — Display character set diagram (page 56)
Fig. 59 · Display character set
Page 57 of the A&D GP-61K user manual — Calibration report using the internal mass
Page 57 · click to enlarge
Page 57

10-2 GLP Report

Set the following parameters to output the report.

z To print the report, set the “GLP output (info)” parameter to “1” and use MODE 3 of the

AD-8121B. For details on using the printer, refer to “15-1 Connection to the AD-8121B Printer”.

z To output the report to a personal computer using the RS-232C interface, set the “GLP output

(info)” parameter to “2”.

z If the time and date are not correct, set the correct time and date in “Clock (Cl adj)” of the

function table.

For operational details about calibration and calibration test, refer to “7. CALIBRATION”.

Calibration report using the internal mass

When the setting is “info 1”: When the setting is “info 2”:

A&D GP-61K — Calibration report using the internal mass diagram (page 57)
Fig. 60 · Calibration report using the internal mass
Page 58 of the A&D GP-61K user manual — Calibration report using an external weight
Page 58 · click to enlarge
Page 58

Calibration report using an external weight

When the setting is “info 1”: When the setting is “info 2”:

A&D GP-61K — Calibration report using an external weight diagram (page 58)
Fig. 61 · Calibration report using an external weight

When the setting is “info 1”: When the setting is “info 2”:

Page 59 of the A&D GP-61K user manual — Operation
Page 59 · click to enlarge
Page 59

Title block and end block

When weight values are recorded as GLP data, a “Title block” is inserted at the beginning and an “End block” is inserted at the end of a group of weight values in the GLP report.

To output the report to an AD-8121B , use MODE 3 of the AD-8121B.

Caution

If the data memory function is used, the “Title block” and “End block” can not be output.

Operation

1 With the weighing data displayed, press and hold the PRINT key, then release the key. S5tart is displayed. The “Title block” is output.

2 The weighing data is output according to the parameter setting of the data output mode.

3 Press and hold the PRINT key until recend is displayed, then release the key. The “End block” is output.

When the setting is “info 1”: When the setting is “info 2”:

A&D GP-61K — Operation diagram (page 59)
Fig. 62 · Operation
Page 60 of the A&D GP-61K user manual — 11-1 Notes on Using Data Memory
Page 60 · click to enlarge
Page 60

11. DATA MEMORY

Data memory is a function to store weighing data, calibration data and unit mass in memory. Of the data in memory, the balance can only display the weighing data. The weighing data and calibration data in memory are available for outputting at one time to a printer or personal computer.

One of the following data sets can be stored:

z Weighing data (Up to 200 sets. 100 sets when the time and date are added.)

z Calibration and calibration test data (latest 50 sets)

z Unit mass in the counting mode (Up to 50 sets)

z Comparator settings (the upper and lower limit values only, up to 20 sets)

z Tare value (Up to 20 sets)

11-1 Notes on Using Data Memory

To use the memory function, set the “Data memory (data)” parameter of the function table. In addition, for weighing data, set the “Time/Date output (5-td)” parameter. For details on setting the data memory, refer to “9. FUNCTION TABLE”.

For weighing data, the data contents to be stored and the storage capacity depend on the “Time/Date output (5-td)” parameter setting.

Releasing “Clr”

If a different type of data exists in memory when the data is stored, “Clr” appears blinking in the upper left of the display. For example, you want to store weighing data but calibration data or unit mass data remains in memory.

Under such a condition, before storing data, delete the data in memory as follows:

1 Press and hold the PRINT key until SClr no with “no” blinking is displayed, then release the key.

2 Press the RE-ZERO key to display SClr go with “go” blinking.

The type of data stored in memory appears in the upper left of the display as shown below:

Unit mass in the counting mode pC Weighing data without time and date -d- Weighing data with time and date d-t Calibration or calibration test data Hi5 Comparator settings Cp Tare value tr

3 Press the PRINT key to delete all the data in memory.

4 The balance displays end and returns to the weighing mode.

Page 61 of the A&D GP-61K user manual — Caution
Page 61 · click to enlarge
Page 61

11-2 Memory for Weighing Data

z The data memory function can store 200 sets of weighing data (100 set if time and date are

added). Even if the AC adapter is removed, the data is maintained in non-volatile memory.

z It is not necessary that the printer or computer be continually connected to the balance,

because the balance stores the weighing data in memory.

z The data in memory is available to be displayed on the balance for confirmation, or to output

several sets of data at one time to a printer or personal computer. In the function setting, what data is to be added to the output data (ID number, data number, time and date) can be selected.

Storing the weighing data

1 Set the “Data memory (data)” parameter to “2”.

2 Set the “Time/Date output (5-td)” parameter as necessary.

3 The storing operation depends on the “Data output mode (prt)” parameter setting. Four types of operating modes are available to store data.

Key mode When the PRINT key is pressed and the displayed value is

stable, the balance stores the weighing data.

Auto print modes A and B When the displayed value is stable and the conditions of “Auto print polarity”, “Auto print difference” and reference value are met, the balance stores the weighing data.

Interval memory mode Weighing data is stored at an interval specified in “Interval time (int)”. Press the PRINT key to start and stop this mode.

A&D GP-61K — Storing the weighing data diagram (page 61)
Fig. 63 · Storing the weighing data

Caution

z When weighing data is being stored in memory, the data can not be output to a personal

computer using the RS-232C interface.

z “fUl” indicates that memory is full or the memory capacity has been reached. More data can

not be stored unless the memory data is deleted.

z Automatic self calibration can not be used while the interval memory mode is active.

Page 62 of the A&D GP-61K user manual — Enabling the data memory function
Page 62 · click to enlarge
Page 62

z The following commands can not be used during data storage.

Q Query command for weighing data.

S Query command for stable weighing data.

SI Query command for weighing data.

SIR Query command for continuous weighing data.

Setting the function table

Parameter settings for each output mode are as follows:

A&D GP-61K — Setting the function table diagram (page 62)
Fig. 64 · Setting the function table

Key mode prt 0 Not used data 2 Auto print mode A prt 1 data 2 Auto print mode B prt 2

Interval memory mode prt 3 Not used data 2 int 0-8

Set each item, depending on the situation, as follows:

A&D GP-61K — Setting the function table diagram (page 62)
Fig. 65 · Setting the function table

Yes d-no 1 Time only 5-td 1 No 5-id 0 Date only 5-td 2 ID number

Both 5-td 3 With 5-td 1, 5-td 2 or 5-td 3 selected, the amount of data to be stored is 100 sets.

Enabling the data memory function

1 Press and hold the SAMPLE key until ba5fnc is displayed, then release the key.

2 Press the SAMPLE key several times to display dout .

3 Press the PRINT key.

4 Press the SAMPLE key three times to display data 0 .

5 Press the RE-ZERO key to display data 2 .

6 Press the PRINT key to store the setting.

7 Press the CAL key to return to the weighing mode.

Page 63 of the A&D GP-61K user manual — Deleting all memory data at one time
Page 63 · click to enlarge
Page 63

Recalling the memory data

Confirm that the “Data memory (data)” parameter is set to “2”.

1 Press and hold the PRINT key until reCall is displayed, then release the key.

2 Press the PRINT key to enter the memory recall mode. The type of data appears in the upper left of the display as shown to the right:

A&D GP-61K — Recalling the memory data diagram (page 63)
Fig. 66 · Recalling the memory data

Recall the data in memory using the following keys

RE-ZERO key To proceed to the next data set.

MODE key To go back to the previous data set.

PRINT key To transmit the current data using the RS-232C interface.

With SAMPLE held down, press the CAL key To delete the data currently displayed.

CAL key To exit the memory recall mode.

3 Press the CAL key to return to the weighing mode.

Transmitting all memory data at one time

Confirm that the “Serial interface (5if)” parameters are set properly. For details, refer to “9. FUNCTION TABLE” and “15. CONNECTION TO PERIPHERAL EQUIPMENT”.

1 Press and hold the PRINT key until reCall is displayed, then release the key.

A&D GP-61K — Transmitting all memory data at one time diagram (page 63)
Fig. 67 · Transmitting all memory data at one time

2 Press the SAMPLE key to display out .

3 Press the PRINT key to display Sout no with “no” blinking.

5 Press the RE-ZERO key to display out go with “go” blinking.

6 Press the PRINT key to transmit all data using the RS-232C interface.

7 The balance displays Clear when all data is transmitted. Press the CAL key to return to the weighing mode.

Deleting all memory data at one time

1 Press and hold the PRINT key until reCall is displayed, then release the key.

2 Press the SAMPLE key several times to display Clear .

3 Press the PRINT key to display SClr no with “no” blinking.

4 Press the RE-ZERO key to display SClr go with “go” blinking.

5 Press the PRINT key to delete all data

6 The balance displays end , then reCall when all data is deleted.

7 Press the CAL key to return to the weighing mode.

Page 64 of the A&D GP-61K user manual — Deleting the memory data
Page 64 · click to enlarge
Page 64

11-3 Memory for Calibration and Calibration Test Data

z Calibration data (when and how it is performed) and calibration test data can be stored in memory.

z All the data in memory is available to be output at one time to a printer or personal computer.

z Up to 50 data sets of the latest calibration or calibration test can be stored. When the memory

capacity has been reached, “fUl” illuminates in the upper left of the display as shown below.

Storing the calibration and calibration test data

If “Clr” appears blinking in the upper left of the display, delete the data in memory.

Store the calibration and calibration test data as follows:

1 Set the “Data memory (data)” parameter to “3”.

2 Set the “GLP output (info)” parameter to “1” or “2”.

3 With the settings above, each time calibration or calibration test is performed, the data is stored automatically.

Transmitting the memory data

Confirm that the “Serial interface (5if)” parameters are set properly. For details, refer to “9. FUNCTION TABLE” and “15. CONNECTION TO PERIPHERAL EQUIPMENT” and also confirm that the “Data memory (data)” parameter is set to “3”.

1 Press and hold the PRINT key until out is displayed, then release the key.

2 Press the PRINT key to display Sout no with “no” blinking.

3 Press the RE-ZERO key to display out go with “go” blinking.

4 Press the PRINT key to transmit all memory data using the RS-232C interface.

5 The balance displays Clear when all memory data is output. Press the CAL key to return to the weighing mode.

Deleting the memory data

1 Press and hold the PRINT key until out is displayed, then release the key.

2 Press the SAMPLE key to display Clear .

3 Press the PRINT key to display SClr no with “no” blinking.

4 Press the RE-ZERO key to display SClr go with “go” blinking.

5 Press the PRINT key to delete all data

6 The balance displays out when all the data has been deleted. Press the CAL key to return to the weighing mode.

Page 65 of the A&D GP-61K user manual — Storing the unit mass
Page 65 · click to enlarge
Page 65

11-4 Memory for Unit Mass in the Counting Mode

z The data memory function can store 50 sets of unit mass for the counting mode. Even if the

AC adapter is removed, the data is maintained in non-volatile memory. Among the 50 sets, “p01” is the memory function for the standard counting mode. z The unit mass in memory can be recalled and used for weighing. z The unit mass in memory can be recalled and changed.

Storing the unit mass

To store a new unit mass: Recall the stored unit mass to be changed (“Pp01” to “p50”). Then, change the

recalled unit mass using the weighing input mode or the digital input mode, and store the new unit mass.

1 Set the “Data memory ( data )” parameter to “1”.

2 Press the MODE key to select pcs (counting mode). If “Clr” appears blinking in the upper left of the display, delete the data in memory.

3 Press and hold the PRINT key until the balance enters the sample unit mass confirmation mode. The unit mass last selected is displayed as shown below.

All the digits blinking

4 Select the unit mass number to be used, using the following keys.

RE-ZERO key To increase the unit mass number by one. MODE key To decrease the unit mass by one.

5 To change the selected unit mass:

To use the weighing input mode, press the SAMPLE key to enter the weighing input mode. Go to “Weighing input mode” on the next page.

To use the digital input mode, press the SAMPLE key, then press and hold the MODE key to enter the digital input mode. Go to “Digital input mode” on the next page.

Using the “UN:mm” command, the unit mass can be recalled. (mm indicates a two-digit numerical value 01-50, which corresponds to P01-P50.) The unit mass recalled can be output using the “?UW” command and can be changed using the “UW:” command.

Page 66 of the A&D GP-61K user manual — Recalling the unit mass
Page 66 · click to enlarge
Page 66

Weighing input mode

In the weighing input mode, the specified number of samples is placed on the pan to store the unit

mass. Re-storing the unit mass or performing Automatic Counting Accuracy Improvement (ACAI) on

the re-stored unit mass is possible. Follow the procedure described in “5-2 Counting Mode (PCS)”. Use the following keys to store a unit mass in weighing input mode.

RE-ZERO key To set the display to zero. 10 - pcs → 10 0 pcs SAMPLE key To change the number of samples to be stored. 10 0 pcs → 25 0 pcs PRINT key Press, after a sample is placed, to store the unit mass. CAL key To return to the unit mass confirmation mode (step 3 in “Recalling the

unit mass”). MODE key Press and hold to go to the digital input mode.

Digital input mode

To use this mode, the sample unit mass must be known beforehand. In the digital input mode, the

unit mass value is entered digitally using the keys.

The display in the digital input mode is shown to the right.

The digit to be changed blinks. Note ACAI can not be used on the unit mass stored using the digital input mode. Use the following keys to store a unit mass in digital input mode.

A&D GP-61K — Digital input mode diagram (page 66)
Fig. 68 · Digital input mode

SAMPLE key To select the digit to be changed. RE-ZERO key To change the value of the

selected digit. MODE key To change the decimal point position. PRINT key To store the unit mass. CAL key To return to the unit mass

confirmation mode (step 3 in “Recalling the unit mass”). MODE key Press and hold to go to the

weighing input mode.

A&D GP-61K — Digital input mode diagram (page 66)
Fig. 69 · Digital input mode

Recalling the unit mass

1 Follow steps1 through 3 in “Storing the unit mass” on the previous page, to enter the sample unit mass confirmation mode. 2 Select the unit mass number to be used, using the following keys.

RE-ZERO key To increase the unit mass

number by one. MODE key To decrease the unit mass by one.

3 Press the PRINT key to confirm the selection and to return to the weighing mode. To cancel the selection and return to the weighing mode, press the CAL key.

Page 67 of the A&D GP-61K user manual — Storing the upper and lower limit values
Page 67 · click to enlarge
Page 67

11-5 Memory for Comparator Settings

z The data memory function can store 20 sets of upper and lower limit values for the comparator mode.

The reference value or tolerance value for the comparator mode can not be stored in memory. z The upper and lower limit values in memory can be recalled easily using the MODE key and

used for weighing. z The upper and lower limit values in memory can be recalled and changed.

Storing the upper and lower limit values

To store new upper and lower limit values: Recall the stored upper and lower limit values to be changed (“C01” to “C20”). Then, change the recalled upper and lower limit values using the digital input mode or the weighing input mode, and store the new value.

1 Press the MODE key to select a unit to be used for storage. Note While the data memory function is in use, unit selection using the MODE key is not available.

2 Set the “Data memory (data)” parameter to “4”. If “Clr” appears blinking in the upper left of the display, delete the data in memory as described in “11-1 Notes on Using Data Memory”.

3 Press and hold the PRINT key until the balance enters the upper and lower limit values confirmation mode. The upper limit value last selected is displayed as shown below.

A&D GP-61K — Storing the upper and lower limit values diagram (page 67)
Fig. 70 · Storing the upper and lower limit values

4 Select the comparator number to be used, using the following keys.

RE-ZERO key To increase the comparator number by one. MODE key To decrease the comparator number by one. Each time the RE-ZERO key or MODE key is pressed, the displayed

value changes as follows: ↔ C03 HI ↔ C03 LO ↔ C04 HI ↔ C04 LO ↔

5 To change the selected upper and lower limit values:

To use the digital input mode, press the SAMPLE key to enter the digital input mode. Go to “Digital input mode” on the next page.

To use the weighing input mode, press the SAMPLE key, then press and hold the MODE key to enter the weighing input mode. Go to “Weighing input mode” on the next page. Note Using the “CN:mm” command, the upper and lower limit values can be recalled. (mm indicates a two-digit numerical value 01-20, which corresponds to C01-C20.) The upper and lower limit values recalled can be output using the “?HI” command or “?LO” command and can be changed using the “HI:” command or “LO:” command.

Page 68 of the A&D GP-61K user manual — Weighing input mode
Page 68 · click to enlarge
Page 68

Digital input mode

In the digital input mode, the upper and lower limit values are entered digitally using the keys.

The display in the digital input mode is shown below.

A&D GP-61K — Digital input mode diagram (page 68)
Fig. 71 · Digital input mode

Use the following keys to store upper and lower limit values in digital input mode.

SAMPLE key To select the digit to be changed. RE-ZERO key To change the value of the

A&D GP-61K — Digital input mode diagram (page 68)
Fig. 72 · Digital input mode

selected digit. MODE key To switch the polarity. PRINT key To store the upper and lower

limit values. CAL key To return to the upper and lower

limit values confirmation mode (step 3 in “Storing the upper and lower limit values”). MODE key Press and hold to go to the

weighing input mode.

Weighing input mode

In the weighing input mode, a sample is placed on the pan to store the upper and lower limit values.

To go to the digital input mode, press and hold the MODE key.

1 The first display in the weighing input mode depends on the comparator number selected in step 4 in “Storing the upper and lower limit values”. For example, when “C03 HI “ is selected in step 4, the display is the current weight value and the comparator number with HI blinking.

A&D GP-61K — Weighing input mode diagram (page 68)
Fig. 73 · Weighing input mode

2 Place a container on the weighing pan, if necessary. Press the RE-ZERO key to set the display to zero.

3 Place a sample corresponding to the upper limit value, on the pan or in the container. 4 Press the PRINT key to store the upper limit value.

Page 69 of the A&D GP-61K user manual — 11-6 Memory for Tare Value
Page 69 · click to enlarge
Page 69

Recalling the upper and lower limit values (Quick selection mode)

The procedure below describes an easy way to recall the upper and lower limit values to be used for

weighing. When the recalled upper and lower limit values are to be changed, refer to “Storing the

upper and lower limit values”.

1 Set the “Data memory (data)” parameter to “4”.

2 Press the MODE key to enter the upper/lower limit value recalling mode. The upper limit value last selected with its comparator number appears. The display is as shown to the right, with all the digits blinking.

A&D GP-61K — Recalling the upper and lower limit values (Quick selection mode) diagram (page 69)
Fig. 74 · Recalling the upper and lower limit values (Quick se…

3 Press the MODE key to select the value. Each time the MODE key is pressed, the displayed value changes as follows: C03 HI → C03 LO→ C04 HI → C04 LO → → C20 LO → C01 HI →.

4 Press the PRINT key to confirm the selection. The balance returns to the weighing mode with the selected upper and lower limit values ready for use.

To cancel the selection, press the CAL key. The balance returns to the weighing mode.

11-6 Memory for Tare Value

z The data memory function can store 20 sets of tare value for weighing.

z The tare value in memory can be recalled easily using the MODE key and used for weighing.

z The tare value in memory can be recalled and changed.

The NET indicator illuminates during tare operation.

Page 70 of the A&D GP-61K user manual — Storing the tare value
Page 70 · click to enlarge
Page 70

Storing the tare value

To store a new tare value: Recall the stored tare value to be changed (“t01” to “t20”). Then, change

the recalled tare value using the digital input mode or the weighing input mode, and store the new

value.

“t-- ” appears when a tare operation is performed without using the tare value stored in memory.

1 Press the MODE key to select a unit to be used for storage.

2 Set the “Data memory (data)” parameter to “5”. If “Clr” appears blinking in the upper left of the display, delete the data in memory as described in “11-1 Notes on Using Data Memory”.

3 Press and hold the PRINT key until the balance enters the tare value confirmation mode. The tare value last selected is displayed as shown below.

A&D GP-61K — Storing the tare value diagram (page 70)
Fig. 75 · Storing the tare value

4 Select the tare number to be used, using the following keys.

RE-ZERO key To increase the tare number by one. MODE key To decrease the tare number by one. Each time the RE-ZERO key or MODE key is pressed, the displayed

value changes as follows: ↔ t03 ↔ t04 ↔ ↔ t20 ↔ t01 ↔ CAL key To cancel the operation and return to the weighing mode.

5 To change the selected tare value:

To use the weighing input mode, press the SAMPLE key to enter the weighing input mode. Go to “Weighing input mode” on the next page.

To use the digital input mode, press the SAMPLE key, then press and hold the MODE key to enter the digital input mode. Go to “Digital input mode” on the next page.

Page 71 of the A&D GP-61K user manual — Digital input mode
Page 71 · click to enlarge
Page 71

Weighing input mode

In the weighing input mode, a sample tare container is placed on the pan to store the tare value.

To go to the digital input mode, press and hold the MODE key.

A&D GP-61K — Weighing input mode diagram (page 71)
Fig. 76 · Weighing input mode

1 The first display in the weighing input mode is the current weight value and the selected tare number with PT blinking.

2 Press the RE-ZERO key to set the display to zero.

3 Place a tare (a container) on the pan.

4 Press the PRINT key to store the tare value. Remove the tare from the pan.

Digital input mode

In the digital input mode, the tare value is entered digitally using the keys.

The display in the digital input mode is shown below.

The digit to be changed blinks.

Use the following keys to store a tare value in digital input mode.

A&D GP-61K — Digital input mode diagram (page 71)
Fig. 77 · Digital input mode

SAMPLE key To select the digit to be changed. RE-ZERO key To change the value of the

selected digit. PRINT key To store the tare value. CAL key To return to the tare value

confirmation mode (step 3 in “Storing the tare value”). MODE key Press and hold to go to the

weighing input mode.

Page 72 of the A&D GP-61K user manual — Canceling the tare value data
Page 72 · click to enlarge
Page 72

Recalling the tare value (Quick selection mode)

The procedure below describes an easy way to recall the tare value to be used for weighing. When

the recalled tare value is to be changed, refer to “Storing the tare value”.

1 Set the “Data memory (data)” parameter to “5”.

2 Press the MODE key to enter the tare value recalling mode. The tare value last selected with its tare number appears. The display is as shown to the right, with all the digits blinking.

A&D GP-61K — Recalling the tare value (Quick selection mode) diagram (page 72)
Fig. 78 · Recalling the tare value (Quick selection mode)

3 Press the MODE key to select the value. Each time the MODE key is pressed, the displayed value changes as follows: t03 → t04 → t05 → → t20 → t01→.

4 Press the PRINT key to confirm the selection. The balance returns to the weighing mode with the selected tare value ready for use.

To cancel the selection, press the CAL key. The balance returns to the weighing mode.

Canceling the tare value data

Cancel the tare value data as follows.

A&D GP-61K — Canceling the tare value data diagram (page 72)
Fig. 79 · Canceling the tare value data

1 Remove everything from the weighing pan and press the RE-ZERO key to cancel the tare value.

Page 73 of the A&D GP-61K user manual — 11-7 Data Memory: Quick Selection Mode
Page 73 · click to enlarge
Page 73

11-7 Data Memory: Quick Selection Mode

The data memory has a quick selection mode, to recall data in memory quickly.

Using the quick selection mode, the comparator settings or the tare value, whichever is selected in

the function table, can be recalled, by a simple operation, using the MODE key.

While the data memory function is in use, unit selection using the MODE key is not available.

Use the quick selection mode as follows.

1 Press the MODE key to enter the memory data recalling mode. The memory data last selected appears with all the digits blinking.

2 Press the MODE key to select the value.

Comparator settings: when the “Data memory (data)” parameter is set to “4” Each time the MODE key is pressed, the displayed

value changes as follows: C03 HI → C03 LO→ C04 HI → C04 LO → → → C20 LO → C01 HI →

A&D GP-61K — 11-7 Data Memory: Quick Selection Mode diagram (page 73)
Fig. 80 · 11-7 Data Memory: Quick Selection Mode

Tare value: when the “Data memory (data)” parameter is set to “5” Each time the MODE key is pressed, the displayed

value changes as follows: t03 → t04 → t05 → → t20 → t01→

3 Press the PRINT key or leave the balance as is for a while (after a few seconds of inactivity) to confirm the selection. The balance returns to the weighing mode with the selected data ready for use.

To cancel the selection, press the CAL key. The balance returns to the weighing mode

Page 74 of the A&D GP-61K user manual — 11-8 Data Memory: Confirmation and Storage Mode
Page 74 · click to enlarge
Page 74

11-8 Data Memory: Confirmation and Storage Mode

The confirmation and storage mode can change and store the recalled data.

Of the three types of memory data listed below, one may be selected in the function table, and is

available for changes using this mode.

z Unit mass in the counting mode z Comparator settings z Tare value

For a detailed description of the procedure for each memory data, see the relevant section.

Use the confirmation and storage mode as follows.

1 Press and hold the PRINT key to enter the memory data recalling mode. The memory data last selected appears with all the digits blinking.

2 Use the following keys to select the value. RE-ZERO key To increase the memory data number by one. MODE key To decrease the memory data number by one.

Unit mass: when the “Data memory (data)” parameter is set to “1” Each time the MODE key or RE-ZERO key is

pressed, the displayed value changes as follows: p03 ↔ p04↔ p05↔ ↔ p20 ↔ p01 ↔

Comparator settings: when the “Data memory (data)” parameter is set to “4” Each time the MODE key or RE-ZERO key is

A&D GP-61K — 11-8 Data Memory: Confirmation and Storage Mode diagram (page 74)
Fig. 81 · 11-8 Data Memory: Confirmation and Storage Mode

pressed, the displayed value changes as follows: C03 HI ↔ C03 LO ↔ C04 HI ↔ C04 LO ↔ ↔ ↔ C20 LO ↔ C01 HI ↔

Tare value: when the “Data memory (data)” parameter is set to “5” Each time the MODE key or RE-ZERO key is

pressed, the displayed value changes as follows: t03 ↔ t04 ↔ t05 ↔ ↔ t20 ↔ t01↔

3 To perform a weighing using the selected memory data Press the PRINT key to confirm the selection. The balance will return to the weighing mode with the selection ready for use. To change and store the selected memory data

Press the SAMPLE key to enter the storage mode. Two inputting modes are available for entering the value to be stored. z Digital input mode z Weighing input mode To switch the inputting mode, press and hold the MODE key. To cancel the operation and return to the weighing mode, press the CAL key.

Page 75 of the A&D GP-61K user manual — Operation
Page 75 · click to enlarge
Page 75

12. PROGRAMMABLE-UNIT

This is a programmable unit conversion function. It multiplies the weighing data in kilograms by an arbitrary coefficient set in the function table and displays the result.

The coefficient must be within the range between the minimum and maximum shown below. If the coefficient set is beyond the range, an error is displayed and the balance returns to the coefficient setting mode, prompting to enter an appropriate value. A coefficient of 1 was set at the factory.

A&D GP-61K — 12. PROGRAMMABLE-UNIT diagram (page 75)
Fig. 82

Model Minimum coefficient Maximum coefficient GP-12K/20K/30K/30KS/40K/61K/61KS 10000 GP-60K/60KS/100K/100KS 1000 GP-32K/32KS 1000 GP-102K

Operation

1 Press and hold the SAMPLE key until ba5fnc of the function table is displayed, then release the key.

2 Press the SAMPLE key several times to display mltCadj .

3 Press the PRINT key. The balance enters the mode to confirm or set the coefficient.

4 The current coefficient is displayed with the first digit blinking. z When it is not to be changed, press the CAL key and proceed to step 6. z When it is to be changed, press the RE-ZERO key

A&D GP-61K — Operation diagram (page 75)
Fig. 83 · Operation

and proceed to step 5.

5 Set the coefficient using the following keys.

SAMPLE key To select a digit to change the

value. The selected digit blinks. RE-ZERO key To change the value. MODE key To change the decimal point

the decimal point position changes as follows: PRINT key To store the new setting, display end and go to step 6. CAL key To cancel the new setting and go to step 6.

6 The balance displays Unit . Press the CAL key to exit the programmable-unit function and return to the weighing mode.

Press the MODE key to select the programmable-unit (no display on the unit section). Perform weighing as described in “5-1 Basic Operation (Kilogram Mode)”. After weighing, the balance displays the result (weighing data in kilograms x coefficient).

Show 4 more labels / short text on this page
Confirming the coefficient · Setting the coefficient · Quitting the operation · Using the function
Page 76 of the A&D GP-61K user manual — 13. DENSITY MEASUREMENT
Page 76 · click to enlarge
Page 76

13. DENSITY MEASUREMENT

The GP series balance is equipped with a density mode. It calculates the density of a solid using the weight of a sample in air and the weight in liquid.

z The density mode was not selected for use when the balance was shipped from the factory. To

use the mode, change the function table and activate the density mode.

z Inputting order, of the parameters necessary for density measurement, can be changed.

z A part of the parameters can be changed and used for re-calculating the density. Press the SAMPLE key to confirm or select each parameter. The parameters are stored in non-volatile memory. Each time a part of the parameters are changed, the density can be re-calculated.

z Two ways to set the density of a liquid are available in the function table, “Liquid density input

(ldin)”: by entering the water temperature or by entering the density directly.

z When the density mode is selected, the response adjustment and self check function can not be used. Formula to obtain the density

The density can be obtained by the following formula.

Where A: Density of a sample A: Weight value of a sample in air B: Weight value of a sample in liquid B : Density of a liquid Prior to measurement: Changing the function table

Prior to measurement, change the function table as follows.

(1) Selecting the density mode

To use the density mode, select it in the function table. Once selected, the density mode is available as one of the weighing units and can be selected by pressing the MODE key. For details, refer to “4-2 Changing the Units”.

(2) Selecting a way to set the density of a liquid

Select the liquid density input method from the function table below. The function table is available only when the density mode is selected.

Class Item Parameter Description 0 Water temperature d5 fnc Density function

ld in Liquid density input 1 Liquid density 0 = Factory setting Measuring the density

(1) Entering the density mode

1 Press the MODE key as necessary to select the density mode. When the density mode begins, “d-a” (weight in air), “g (kg for the GP-102K)” for the unit and the processing indicator are turned on.

Page 77 of the A&D GP-61K user manual — 9-11 Adding the Comparison Results
Page 77 · click to enlarge
Page 77

(2) Selecting a parameter to set.

A&D GP-61K — 9-11 Adding the Comparison Results diagram (page 77)
Fig. 84 · 9-11 Adding the Comparison Results

1 Press the SAMPLE key to select a parameter to set. Each pressing of the key switches the parameter. 2 Press the RE-ZERO key to enter the inputting mode of the parameter selected. (3) Setting the parameter.

A. Entering the weight of a sample in air.

1 Confirm that nothing is placed on the upper pan (in air). Press the RE-ZERO key to set the display to zero. 2 Place a sample on the upper pan. When the value displayed on the balance becomes stable, press the PRINT key. The sample weight in air is confirmed and the balance will return to the density mode (the condition of step 1, (2) Selecting a parameter to set). 3 Remove the sample from the upper pan. Notes If an unstable value, a negative value or e (error) is displayed, the PRINT key is disabled.

The SAMPLE key can not be used to change the minimum weighing value.

B. Entering the weight of a sample in liquid.

1 Confirm that nothing is placed on the lower pan (in liquid). Press the RE-ZERO key to set the display to zero. 2 Place a sample on the lower pan (in liquid). When the value displayed on the balance becomes stable, press the PRINT key. The sample weight in liquid is confirmed and the balance will return to the density mode (the condition of step1, (2) Selecting a parameter to set). 3 Remove the sample from the lower pan. Notes If an unstable value or e (error) is displayed, the PRINT key is disabled.

The SAMPLE key can not be used to change the minimum weighing value.

Page 78 of the A&D GP-61K user manual — 9-11 Adding the Comparison Results
Page 78 · click to enlarge
Page 78

C. Entering the density of a liquid.

Two ways to set the density of a liquid are available in the function table, Liquid density input (ldin): by entering the water temperature or by entering the density directly.

  • Enter the water temperature ( ldin 0 ) The water temperature currently set (unit: ° C, factory setting : 25 ° C) is displayed.

Use the following keys to change the value.

RE-ZERO key To increase the temperature by one degree. (0-99°C)

MODE key To decrease the temperature by one degree. (0-99°C)

PRINT key To store the change, display end and return to the density mode (the condition of step1, (2) Selecting a parameter to set).

CAL key To cancel the change and return to the density mode (the condition of step 1, (2) Selecting a parameter to set).

The relation between the water temperature and density is shown below.

A&D GP-61K — 9-11 Adding the Comparison Results diagram (page 78)
Fig. 85 · 9-11 Adding the Comparison Results
  • Enter the density directly ( ldin 1 ) The density currently set (unit : g / cm 3 , factory setting : 1.0000g / cm 3 ) is displayed.

Use the following keys to change the value.

RE-ZERO key To set the value of the digit selected.

SAMPLE key To select the digit to change the value.

PRINT key To store the change, display end and return to the density mode (the condition of step1, (2) Selecting a parameter to set).

CAL key To cancel the change and return to the density mode (the condition of step1, (2) Selecting a parameter to set).

(4) Outputting the density.

After all the necessary parameters have been set, press the SAMPLE key to output the density calculated.

Page 79 of the A&D GP-61K user manual — RS-232C
Page 79 · click to enlarge
Page 79

14. I/O UNIT SPECIFICATIONS (Standard)

14-1 RS-232C//External Contact Input

D-Sub 25 pin numbers

D-Sub 25 pin assignments Pin No. Signal name Interface type Direction Description 1 FG - Frame ground 2 RXD RS-232C Input Receive data 3 TXD RS-232C Output Transmit data 4 RTS RS-232C Input Ready to send 5 CTS RS-232C Output Clear to send 6 DSR RS-232C Output Data set ready 7 SG RS-232C/external contact input - Signal ground 18 PRINT External contact input Input Same as the PRINT key 19 RE-ZERO External contact input Input Same as the RE-ZERO key Others - - - No connection

A&D GP-61K — D-Sub 25 pin numbers diagram (page 79)
Fig. 86 · D-Sub 25 pin numbers

RS-232C

The balance is a DCE device. Connect the balance to a personal computer (DTE), using a straight

through cable.

Transmission system : EIA RS-232C Transmission form : Asynchronous, bi-directional, half duplex Transmission rate : 10 times/second or 5 times/second (same as data refresh rate) Data format : Baud rate : 600, 1200, 2400, 4800, 9600, 19200bps Data bits : 7 or 8 bits Parity : Even, Odd (Data bits 7 bits) None (Data bits 8 bits) Stop bit : 1 bit

(When sending, 2 bits; receiving, 1 bit. A personal computer will function with either

setting.) CodeASCII
A&D GP-61K — RS-232C diagram (page 79)
Fig. 87
Page 80 of the A&D GP-61K user manual — 9-11 Adding the Comparison Results
Page 80 · click to enlarge
Page 80
A&D GP-61K — 9-11 Adding the Comparison Results diagram (page 80)
Fig. 88 · 9-11 Adding the Comparison Results

By connecting pin 18 to pin 7, or pin 19 to pin 7, for more than 100 ms, the same operation as performed by pressing the RE-ZERO key or the PRINT key, will be performed. Use of example

A&D GP-61K — 9-11 Adding the Comparison Results diagram (page 80)
Fig. 89 · 9-11 Adding the Comparison Results
A&D GP-61K — 9-11 Adding the Comparison Results diagram (page 80)
Fig. 90 · 9-11 Adding the Comparison Results

The external input connector (AX-HDB-25P/CTF) and the foot switch (AX-SW128) are sold separately.

Page 81 of the A&D GP-61K user manual — 15-2 Connection to a Computer
Page 81 · click to enlarge
Page 81

15. CONNECTION TO PERIPHERAL EQUIPMENT

15-1 Connection to the AD-8121B Printer

Set the following parameters to use the AD-8121B printer.

Function setting Description dout prt 0-3 Selects a print mode. dout ap-p 0-2 Selects the polarity for the auto print mode. dout ap-b 0-2 Selects the auto print difference. dout pU5e 0,1 Selects data output pause. 5if bp5 2 Factory setting 2400 bps 5if btpr 0 Factory setting 7 bits, Even parity check 5if Crlf 0 Factory setting CR, LF 5if Ct5 0 Factory setting CTS and RTS control, not used

A&D GP-61K — 15-1 Connection to the AD-8121B Printer diagram (page 81)
Fig. 91 · 15-1 Connection to the AD-8121B Printer

When “MODE 1” or “MODE 2” of the AD-8121B printer is used dout 5-td 0 Factory setting Not to output the time and date. dout 5-id 0 Factory setting Not to output the ID number. dout at-f 0 Factory setting Not to use auto feed. 5if type 0 Factory setting A&D standard format

A&D GP-61K — 15-1 Connection to the AD-8121B Printer diagram (page 81)
Fig. 92 · 15-1 Connection to the AD-8121B Printer

When “MODE 3” of the AD-8121B printer is used. dout 5-td 0-3 Outputs the time and date as necessary. dout 5-id 0,1 Outputs the ID number as necessary. dout at-f 0,1 Uses auto feed as necessary. 5if type 1 DP format

A&D GP-61K — 15-1 Connection to the AD-8121B Printer diagram (page 81)
Fig. 93 · 15-1 Connection to the AD-8121B Printer

When data is transmitted continuously. When all memory data is transmitted at one time. dout pU5e 1 Uses pause.

A&D GP-61K — 15-1 Connection to the AD-8121B Printer diagram (page 81)
Fig. 94 · 15-1 Connection to the AD-8121B Printer

Setting What is printed data 0 The weighing data data 2 The weighing data stored in memory data 3 The calibration or calibration test data stored in memory

Refer to “10. ID NUMBER AND GLP REPORT” for a print sample.

15-2 Connection to a Computer

The GP series balance is of the DCE type (Data Communication Equipment), which can be connected to a personal computer using the RS-232C interface. Before connection, read the personal computer manual thoroughly. Use a standard DCE cable for connection (cable type: straight-through). When the personal computer type is a DOS/V with a 9-pin port, use a straight-through cable with a 25-pin male connector and a 9-pin female connector.

Page 82 of the A&D GP-61K user manual — 15-4 Using the WinCT software, the balance can do the following:
Page 82 · click to enlarge
Page 82

15-3 Using Windows Communication Tools (WinCT)

When Windows is used as an operating system in a personal computer, the provided WinCT software can be used to transmit the weighing data to the personal computer. The WinCT software has two communication methods: “RsCom” and “RsKey”. For details on WinCT, refer to the WinCT instruction manual. RsCom

z Can transmit commands to control the balance.

z Can make bi-directional communication between the balance and a personal computer using

the RS-232C interface.

z Can display or store the data using a text file format. Can also print the data using a printer

connected to the personal computer.

z When several ports of a personal computer have balances connected, can communicate with

each balance simultaneously.

z Can share a personal computer with other application software.

z Can receive the balance GLP report. RsKey

z Can transmit the weighing data output from the balance directly to other application software

such as Microsoft Excel.

z Can be used with most application software.

z Can receive the balance GLP report.

15-4 Using the WinCT software, the balance can do the following:

1 Analyzing the weighing data and the statistics with “RsKey” The weighing data can be input directly into an Excel worksheet. Then, Excel can analyze the data to obtain total, average, standard deviation, maximum and minimum value, and display them in a graph. 2 Controlling the balance using commands from a personal computer By using “RsCom”, the personal computer sends commands such as “re-zero” or “send weighing data” to the balance and controls the balance.

3 Printing the balance GLP report using your printer The balance GLP report can be printed using a printer connected to the personal computer.

4 Receiving weighing data at a certain interval The weighing data can be received at a certain interval and data characteristic with elapsed time can be obtained.

5 Using the GP series balance memory function The data can be stored in the balance’s memory. Of the data stored, the weighing data and calibration data can be transmitted to a personal computer at one time.

6 Using a personal computer as an external indicator With the “RsKey” test mode function, a personal computer can be used as an external weight indicator for the balance. (To do this, set the balance data output mode to stream mode.)

Page 83 of the A&D GP-61K user manual — 16-1 Command List
Page 83 · click to enlarge
Page 83

16. COMMANDS

16-1 Command List

A command has a terminator added, that is specified using “5if Crlf” of the function table, and is sent to the balance.

Commands to query weighing data C Cancels the S or SIR command. Q Requests the weighing data immediately. S Requests the weighing data when stabilized. SI Requests the weighing data immediately. SIR Requests the weighing data continuously. Commands to control the balance CAL Same as the CAL key. OFF Turns the display off. ON Turns the display on. P Same as the ON:OFF key PRT Same as the PRINT key R Same as the RE-ZERO key SMP Same as the SAMPLE key. U Same as the MODE key UN:mm Recalls the unit mass values in memory. (mm indicates 01-50) ?UN Requests the unit mass numbers in memory. UW:∗∗∗∗∗∗. ∗ g Changes the unit mass value. (Only “g” is available as a unit.) Command e.g.: UW:+002000.0 g (Set the unit mass value to 2000.0 g: indicates a space.) ?UW Requests the unit mass value. CN:mm Recalls the upper/lower limit value in memory (mm indicates 01-20) ?CN Requests the upper/lower limit value code number of the selected value. H I:∗∗∗∗∗∗. ∗ g Sets the upper limit values. Command e.g.: HI:+002000.0 g (To set the upper limit value to 2000.0 g: indicates a space.) LO:∗∗∗∗∗∗. ∗ g Sets the lower limit values. Command e.g.:LO:+001000.0 g (To set the lower limit value to 1000.0 g: indicates a space.) ?HI Requests the upper limit value. ?LO Requests the lower limit value. PN:mm Recalls the tare value in memory. (mm indicates 01-20) ?PN Request the tare number of the selected value. (mm indicates 01-20) PT:∗∗∗∗∗. ∗ g Sets the tare value. Command e.g.:PT:+001000.0 g (To set the tare value to 1000.0 g: indicates a space.) ?PT Request the tare value. MCL Deletes all data in memory. MD:nnn Deletes data with the data number nnn. ?MA Outputs all weighing data in memory. ?MQnnn Outputs data with the data number nnn. ?MX Outputs the number of data in memory (the last data number)

When a unit is required in commands such as a “PT:” command, use the 3-digit unit code of the A&D standard format.

nnn indicates a three-digit numerical value.

Page 84 of the A&D GP-61K user manual — 16-4 Settings Related to RS-232C
Page 84 · click to enlarge
Page 84

16-2 Acknowledge Code and Error Codes

When the “Serial interface function (5if)” parameter is set to “erCd 1”, the balance outputs

code or error code to each command as follows:

(06h) Acknowledge in ASCII code.

z When the balance receives a command to request data and can not process it, the balance

transmits an error code (EC, Exx). When the balance receives a command to request data and can process it, the balance outputs the data.

z When the balance receives a command to control the balance and can not process it, the

balance transmits an error code (EC, Exx). When the balance receives a command to control the balance and can process it, the balance transmits the acknowledge code.

Among commands to control the balance, the following transmit the acknowledge code both when the balance receives the command and when the balance has accomplished the command. If the command can not be processed properly, the balance transmits an error code (EC, Exx). This error can be released using the CAL command.

CAL command (Calibration command) ON command (Display ON command)

P command (Display ON/OFF command) R command (RE-ZERO command)

z When a communication error has occurred due to external noise, or a parity error has occurred

due to transmission error, the balance transmits an error code. In this case, send the command again.

16-3 Control Using CTS and RTS

Depending on the “Ct5” parameter of “Serial interface (5if)”, the balance performs as follows:

Regardless of whether the balance can receive a command or not, the balance keeps the CTS line HI. The balance outputs data regardless of the condition of the RTS line.

The CTS line is kept Hi normally. When the balance can not receive the next command (e.g. while the balance is processing last command), the balance sets the CTS line to Lo. The balance confirms the level of the RTS line before outputting a set of data. If the RTS level is Hi, the balance outputs data. If the RTS level is Lo, data is not output (The data is canceled).

16-4 Settings Related to RS-232C

Concerning the RS-232C, the balance has two functions: “Data output (dout)” and “Serial interface (5if)”. Set each function as necessary.

Ct5 0 · Ct5 1

Page 85 of the A&D GP-61K user manual — Checking that the operating environment or weighing method is proper
Page 85 · click to enlarge
Page 85

17. MAINTENANCE

z Do not disassemble the balance. Contact the local A&D dealer if the balance needs service or repair.

z Use the original packing material for transportation.

z Do not use organic solvents to clean the balance. Clean the balance with a lint free cloth that

is moistened with warm water and a mild detergent.

z The mass of the internal mass may change due to corrosion or other damage caused by the operating

environment, or due to aging. Check the internal mass using an external weight periodically.

18. TROUBLESHOOTING

18-1 Checking the Balance Performance and Environment

The balance is a precision instrument. When the operating environment or the operating method is

inadequate, correct weighing can not be performed. Place a sample on the pan and remove it, and

repeat this several times. If the balance seems to have a problem with repeatability or to perform

improperly, check as described below. If improper performance persists after checking, contact the

local A&D dealer for repair.

Checking that the balance performs properly

z Check the balance performance using the self-check function as described in “6-1 Automatic

Response Adjustment / Self Check Function”.

An error display appears when a malfunction is found.

z Check the balance repeatability using an external weight. Be sure to place the weight in the

center of the weighing pan.

z Check the balance repeatability, linearity and calibrated value using external weights with a

known value.

Checking that the operating environment or weighing method is proper

z Is the weighing table solid enough?

z Is the balance level? Refer to “3-1 Before Use”.

z Is the operating environment free from vibration and drafts?

z Is there a strong electrical or magnetic noise source such as a motor near the balance?

z Is the weighing pan installed correctly?

z Is the RE-ZERO key pressed before placing a sample on the weighing pan?

z Is the sample placed in the center of the weighing pan?

z Has the balance been calibrated using the internal mass (one-touch calibration)?

z Has the balance been warmed up for 30 minutes before weighing?

Operating environment · Weighing method

Page 86 of the A&D GP-61K user manual — 18-2 Error Codes
Page 86 · click to enlarge
Page 86

z Has the sample absorbed or lost moisture due to the ambient conditions such as temperature

z Has the temperature of the container been allowed to equalize to the ambient temperature?

Refer to “3-2 During Use”.

z Is the sample charged with static electricity? Refer to “3-2 During Use”.

z Is the sample of magnetic material such as iron? Caution is required for weighing magnetic

materials. Refer to “3-2 During Use”.

18-2 Error Codes

Display Error code Description

EC, E11 Stability error The balance can not stabilize due to an environmental problem. Prevent vibration, drafts, temperature changes, static electricity and magnetic fields, from influencing the balance. Refer to “3. PRECAUTIONS” for details on the operating environment and “6. RESPONSE ADJUSTMENT/SELF CHECK FUNCTION” about adapting the balance to the environment. To return to the weighing mode, press the CAL key.

Out of the setting range The data to be stored is out of the setting range.

EC, E16 Internal mass error Applying the internal mass does not yield a change in the weight value as specified. Confirm that there is nothing on the pan and perform the weighing operation from the beginning again.

EC, E17 Internal mass error The internal mass application mechanism does not function properly. Perform the weighing operation from the beginning again.

EC, E20 Calibration weight error The calibration weight is too heavy. Confirm the calibration weight value. Press the CAL key to return to the weighing mode.

EC, E21 Calibration weight error The calibration weight is too light. Confirm the calibration weight value. Press the CAL key to return to the weighing mode.

Overload error A sample beyond the balance weighing capacity has been placed on the pan. Remove the sample from the pan.

Sample and container · and humidity?

Page 87 of the A&D GP-61K user manual — (CHECK NO)
Page 87 · click to enlarge
Page 87

Display Error code Description

Weighing pan Error The weight value is too light. Confirm that the weighing pan is properly installed and calibrate the balance.

Sample mass error The balance can not store the sample for the counting mode or for the percent mode because it is too light. Use a larger sample.

Unit mass error The sample unit mass for the counting mode is too light. Storing and using it for counting will cause a counting error. Add samples to reach the specified number and press the PRINT key. Pressing the PRINT key without adding samples will shift the balance to the counting mode. But, to acquire accurate weighing, be sure to add samples.

Automatic response adjustment zero error The automatic response adjustment can not be performed because there is something on the pan. Clear the pan. Press the CAL key to return to the weighing mode.

Automatic response adjustment unstable error The automatic response adjustment can not be performed because the weight value is unstable. Check the ambient conditions such as breeze, vibration and magnetic fields, also check the weighing pan. Press the CAL key to return to the weighing mode.

(CHECK NG)

Internal error Indicates an internal error as the result of self-check function. Repair is required. Contact the local A&D dealer.

(CHECK NO)

Clock battery error The clock backup battery has been depleted. Press any key and set the time and date. The clock and calendar function works normally as long as the AC adapter is connected to the balance. If this error appears frequently, contact the local A&D dealer.

Memory full The amount of weighing data in memory has reached the maximum capacity. Delete data in memory to store new data. For details, refer to “11. DATA MEMORY”.

Memory full The amount of calibration or calibration test data in memory has reached the maximum capacity (50 sets). The data in memory will be deleted automatically to store new data. For details, refer to “11. DATA MEMORY”.

(Blinking) · (Illuminated)

Page 88 of the A&D GP-61K user manual — 18-3 Asking For Repair
Page 88 · click to enlarge
Page 88

Display Error code Description

Memory type error Type of memory set in the function table and type of data stored are different. For details, refer to “11. DATA MEMORY”.

EC, E00 Communications error A protocol error occurred in communications. Confirm the format, baud rate and parity.

EC, E01 Undefined command error An undefined command was received. Confirm the command.

EC, E02 Not ready A received command can not be processed. e.g. The balance received a Q command, but not in the

weighing mode. e.g. The balance received a Q command while

processing a RE-ZERO command. Adjust the delay time to transmit a command.

EC, E03 Timeout error If the timeout parameter is set to “t-Up1”, the balance did not receive the next character of a command within the time limit of one second. Confirm the communication.

EC, E04 Excess characters error The balance received excessive characters in a command. Confirm the command.

EC, E06 Format error A command includes incorrect data. e.g. The data is numerically incorrect. Confirm the command.

EC, E07 Parameter setting error The received data exceeds the range that the balance can accept. Confirm the parameter range of the command.

Other errors If the errors described above can not be released or other errors are displayed, contact the local A&D dealer. Other symbol

When this indicator blinks, automatic self calibration is required. The indicator blinks when the balance detects changes in ambient temperature. If the balance is not used for several minutes with this indicator blinking, the balance performs automatic self calibration. The blinking period depends on the operating environment.

18-3 Asking For Repair

If the balance needs service or repair, contact the local A&D dealer.

The balance is a precision instrument. Use much care when handling the balance and observe the following when transporting the balance.

z Use the original packing material.

z Remove the weighing pan from the main unit.

Page 89 of the A&D GP-61K user manual — GP-12K GP-20K GP-30K
Page 89 · click to enlarge
Page 89

19. SPECIFICATIONS

GP-12K GP-20K GP-30K

GP-32K

GP-40K

GP-30KS

GP-32KS

Weighing capacity 12 kg 21 kg 31 kg 31 kg 41 kg

12.0084 kg 21.0084 kg 31.0084 kg 31.008 kg

41.0080 kg

(See bottom for the description of ∗)

6.1009 kg∗

Minimum weighing value (1 digit) 0.1 g 1 g/0.1 g 0.5 g

Repeatability (Standard deviation) 0.1 g 0.5 g/0.1 g 0.5 g

Linearity ±0.2 g ±1g/±0.2 g ±1g

Stabilization time (typical at FAST) Approx.1.5 seconds

Sensitivity drift (10°C-30°C/50°F-86°F ) ∗∗ ±3 ppm/°C ±5 ppm/°C

±1.0 g ±1.5 g Accuracy right after calibration using the internal mass. ∗∗∗ The value above is to the weighing capacity.

Operating environment 5°C to 40°C (41°F to 104°F)

85%RH or less (No condensation)

Display refresh rate 5 times/second or 10 times/second

Minimum unit mass 0.1 g 1 g 0.5 g Counting mode Number of samples 10, 25, 50 or 100 pieces

Minimum 100% reference mass 10 g 100 g 50 g Percent mode Minimum 100% display 0.01%, 0.1%, 1% (Depends on the reference mass stored.)

Interface (Provided as standard) RS-232C with Windows Communication Tools WinCT

External calibration weight 5 kg

Weighing pan 384 x 344 mm

Dust and water protection Complying with IP65

Sensor used Super Hybrid Sensor (SHS)

AC adapter Confirm that the adapter type is correct for the local voltage and power receptacle type

Power consumption Approx. 14VA (supplied to the AC adapter )

External dimensions (WxDxH, mm) 372 x 615 x 130 (GP-30KS/ GP-32KS: 344 x 442 x 130 ) ∗∗∗∗

Net weight Approx. 17 kg

∗ The GP-32K and GP-32KS allow weighing using a precision range, even with a heavy tare

placed on the pan. (Smart range function) ∗∗ When automatic self calibration is not used, at 10°C to 30°C (50°F to 86°F) ∗∗∗ Accuracy right after calibration using the internal mass in good ambient conditions (within the

temperature range of 10°C to 30°C (50°F to 86°F) with no abrupt changes in temperature or humidity, no drafts, no effect by magnetic fields or static electricity). Check the internal mass periodically as described in “17. MAINTENANCE”. ∗∗∗∗ The size for the GP-30KS/32KS is that of the main unit. The display size is 260x164x202. The cable length, for the separate display type balances (GP-30KS/32KS), is 3 meters.

Show 9 more labels / short text on this page
Maximum display · 10 kg · 20 kg · 20 kg · 10 kg · 20 kg · 30 kg · 30 kg · 40 kg
Page 90 of the A&D GP-61K user manual — GP-60K
Page 90 · click to enlarge
Page 90

GP-60K

GP-61K

GP-100K

GP-102K

GP-60KS

GP-61KS

GP-100KS

Weighing capacity 61 kg 61 kg 101 kg 101 kg

Maximum display (See bottom for ∗) 61.084 kg 61.0084 kg 101.084 kg 101.08kg

61.009 kg∗

Minimum weighing value (1 digit) 1 g 0.1 g 1 g 10 g/1 g

Repeatability (Standard deviation) 0.7 g 0.2 g 1 g 5 g/1 g

Linearity ±1 g ±0.5 g ±2 g ±10 g/±2 g

Stabilization time (typical at FAST) Approx.1.5 seconds

Sensitivity drift (10°C-30°C/50°F-86°F ) ∗∗ ±6ppm/°C ±10 ppm/°C

±5 g ±3 g ±10 g Accuracy right after calibration using the internal mass. ∗∗∗ The value above is to the weighing capacity.

Operating environment 5°C to 40°C (41°F to 104°F)

85%RH or less (No condensation)

Display refresh rate 5 times/second or 10 times/second

Minimum unit mass 1 g 0.1 g 1 g 10 g Counting mode Number of samples 10, 25, 50 or 100 pieces

Minimum 100% reference mass 100 g 10 g 100 g 1000 g Percent mode Minimum 100% display 0.01%, 0.1%, 1% (Depends on the reference mass stored.)

Interface (Provided as standard) RS-232C with Windows Communication Tools WinCT

External calibration weight 30 kg 40 kg 50 kg 60 kg

60 kg 80 kg 100 kg

Weighing pan 384 x 344 mm 386 x 346mm

Dust and water protection Complying with IP65

Sensor used Super Hybrid Sensor (SHS)

AC adapter Confirm that the adapter type is correct for the local voltage and power receptacle type

Power consumption Approx. 14VA (supplied to the AC adapter )

External dimensions (WxDxH, mm) 372 x 615 x 130 (GP-60KS/ GP-61KS: 344 x 442 x 130 ) ∗∗∗∗

373 x 615 x 130 (GP-100KS: 346 x 443 x 130 ) ∗∗∗∗

Net weight Approx. 17 kg Approx. 18 kg

∗ The GP-102K allow weighing using a precision range, even with a heavy tare placed on the

pan. (Smart range function) ∗∗ When automatic self calibration is not used, at 10°C to 30°C (50°F to 86°F) ∗∗∗ Accuracy right after calibration using the internal mass in good ambient conditions (within the

temperature range of 10°C to 30°C (50°F to 86°F) with no abrupt changes in temperature or humidity, no drafts, no effect by magnetic fields or static electricity). Check the internal mass periodically as described in “17. MAINTENANCE”. ∗∗∗∗ The size for the GP-60KS/61KS/100KS is that of the main unit. The display size is 260x164x202. The cable length, for the separate display type balances (GP-60KS/61KS /100KS), is 3 meters.

Page 91 of the A&D GP-61K user manual — Standard Type: GP-12K/20K/30K/32K/40K/60K/61K/100K/102K
Page 91 · click to enlarge
Page 91

20. EXTERNAL DIMENSIONS

Standard Type: GP-12K/20K/30K/32K/40K/60K/61K/100K/102K

A&D GP-61K — Standard Type: GP-12K/20K/30K/32K/40K/60K/61K/100K/102K diagram (page 91)
Fig. 95 · Standard Type: GP-12K/20K/30K/32K/40K/60K/61K/100K/1…
A&D GP-61K — Standard Type: GP-12K/20K/30K/32K/40K/60K/61K/100K/102K diagram (page 91)
Fig. 96 · Standard Type: GP-12K/20K/30K/32K/40K/60K/61K/100K/1…
Page 92 of the A&D GP-61K user manual — Separate Display Type: GP-30KS/32KS/60KS/61KS/100KS
Page 92 · click to enlarge
Page 92

Separate Display Type: GP-30KS/32KS/60KS/61KS/100KS

A&D GP-61K — Separate Display Type: GP-30KS/32KS/60KS/61KS/100KS diagram (page 92)
Fig. 97 · Separate Display Type: GP-30KS/32KS/60KS/61KS/100KS

The size in parentheses is for GP-100KS Unit: mm

Page 93 of the A&D GP-61K user manual — 21. OPTIONS
Page 93 · click to enlarge
Page 93

21. OPTIONS

A&D GP-61K — 21. OPTIONS diagram (page 93)
Fig. 98

z Compact dot-matrix printer z Statistical function, clock and calendar function, interval

print function, graphic print function, dump print mode z 5 x 7 dots, 16 characters per line z Print paper (AX-PP143, 45 (W) x 50 (L) mm , ø65 mm) z AC adapter or alkaline battery OP-04: Comparator output (relay/ with a buzzer)/RS-232C/Current loop

z Generates a relay output corresponding to HI , OK and LO in the display. z Current loop and RS-232C can be used at the same time. For example, a personal computer

and an AD-8121B printer can be used simultaneously. z Installed in place of the standard RS-232C serial interface. OP-06: Analog voltage output/Current loop

z Gains a voltage of 0-1V (or 0.2-1V) depending on the displayed value. z With this option, current loop can be used. For example, an AD-8121B printer can be used

simultaneously using this option. z Installed in place of the standard RS-232C serial interface. So, RS-232C is not available for use. OP-07: Extension cable, 5 m (Available as factory option only when ordered with a GP balance)

z Applicable balance models: GP-30KS/32KS/60KS/61KS/100KS z A cable of 5 meters, to connect between the display and the main unit OP-20: Underhook

z Applicable balance models: GP-12K/20K/30K/30KS/32K/32KS/40K z Used for measuring the density of magnetic materials and concrete (coarse aggregates). OP-21: Underhook

z Applicable balance models: GP-60K/60KS/61K/61KS/100K/100KS/102K z Used for measuring the density of magnetic materials and concrete (coarse aggregates). OP-22: Printer bracket

z Used to place the AD-8121B printer beside the display. AD-1682 Rechargeable battery

z Allows use of the balance in a place where AC power is not available. AD-8524A/B: Keyboard adapter

z Used to connect the balance to a personal computer with appropriate OS and applications. AD-8920: Remote display

z Connected to the GP series using the RS-232C interface. AX-SW128: Foot switch

z Used to externally transmit a RE-ZERO or PRINT signal to the balance.

AD-8121B Printer

Page 94 of the A&D GP-61K user manual — Terms
Page 94 · click to enlarge
Page 94

22. TERMS/INDEX

Terms

Stable value The stable weight data, indicated by the illuminated stabilization indicator.

Environment Ambient conditions such as vibration, drafts, temperature, static electricity

and magnetic fields which affect the weighing operation.

Store To save the weighing data, unit mass or calibration data using the data

memory function.

Calibration Adjustment of the balance so that it can weigh accurately.

Output To output the weighing data using the RS-232C interface.

Zero point A weighing reference point. Usually refers to the value displayed when

nothing is on the weighing pan.

Smart range function Allows weighing in the precision range, even if a heavy tare is used.

Data number Numbers assigned sequentially when weighing data or unit weight is stored.

Digit The minimum weighing value available. Used for the balance, 1 digit is the

smallest displayable mass.

Tare To cancel the weight of a container which is not to be included in the

weighing data. Normally, refers to an operation of placing a container and

setting the display to zero.

Mode Balance operational function.

Re-zero To set the display to zero.

GLP Good Laboratory Practice.

Repeatability Variation in measured values obtained when the same mass is placed and

removed repetitively. Usually expressed as a standard deviation.

e.g. Standard deviation=1 digitThis means that measured values, obtained

when the same sample is placed and removed repetitively, fall within ±1 digit

in the frequency of about 68%.

Stabilization time Time required after a sample being placed, until the stabilization indicator

illuminates and the weighing data is displayed.

Sensitivity drift An affect that a change in temperature causes to the weighing data.

Expressed as temperature coefficient.

e.g. Temperature coefficient = 3 ppm/°C : If a load is 20 kg and the

temperature changes by 10°C, the value displayed changes by the following

value.

0.0003%/°C x 10°C x 20 kg = 0.6 g

In this example, if the value displayed is 20.0000 kg before temperature

changes, a temperature change of 10°C will make the value displayed

20.0006 kg.

Page 95 of the A&D GP-61K user manual — 9-11 Adding the Comparison Results
Page 95 · click to enlarge
Page 95
Page 95 of the A&D GP-61K user manual — 9-11 Adding the Comparison Results
Page 96 of the A&D GP-61K user manual — 9-11 Adding the Comparison Results
Page 96 · click to enlarge
Page 96
Page 96 of the A&D GP-61K user manual — 9-11 Adding the Comparison Results
Page 97 of the A&D GP-61K user manual — MEMO
Page 97 · click to enlarge
Page 97

MEMO

Page 98 of the A&D GP-61K user manual — MEMO
Page 98 · click to enlarge
Page 98

MEMO

Display codes

Error codes

What the fault codes shown on the display mean and how to clear them.

CodeMeaningFix
EC, E00Communications error A protocol error occurred in communications. Confirm the format, baud rate and parity. p.88
EC, E01Undefined command error An undefined command was received. Confirm the command. p.88
EC, E02Not ready A received command can not be processed. e.g. The balance received a Q command, but not in the weighing mode. e.g. The balance received a Q command while processing a RE-ZERO command. Adjust the delay time to transmit a command. p.88
EC, E03Timeout error If the timeout parameter is set to “t-Up1”, the balance did not receive the next character of a command within the time limit of one second. Confirm the communication. p.88
EC, E04Excess characters error The balance received excessive characters in a command. Confirm the command. p.88
EC, E06Format error A command includes incorrect data. e.g. The data is numerically incorrect. Confirm the command. p.88
EC, E07Parameter setting error The received data exceeds the range that the balance can accept. Confirm the parameter range of the command. p.88

Ask about the A&D GP-61K

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the A&D GP-61K

Be the first to ask — other owners or our team will answer.