1750 Operators Manualp.1LIMITED WARRANTY AND LIMITATION OF LIABILITYp.2Table of Contentsp.3List of Tablesp.5List of Figuresp.7Introductionp.9Contacting Flukep.10Safety Informationp.11Accessoriesp.12Featuresp.13Applying the Front Panel Decalp.15Charging the PDA Batteryp.17Installing the Softwarep.17Installing Power View on the PDAp.17Language Selectionp.18PC System Requirement for Power Analyzep.18Installing Power Analyzep.18Installing the Power Recorder at a Facilityp.19Work Flowp.19Installing the Recorderp.20Connecting the Recorder to the Wiringp.20Verifying Your Connectionp.21Finishing UPp.21Managing Recorded Datap.22Power Type Diagramsp.22Communicating with the Recorder Using Power Viewp.33Navigating in Power View on the PDAp.34Power View Menusp.34Icons on the Menu Barp.35Menu Treep.36Home Screenp.37Downloading Data and Erasing Memoryp.39Viewing Live Datap.43Scope Screenp.43Meter Screenp.44Phasor Screenp.44Harmonics Screenp.45Trend Screenp.45Viewing Phasesp.46Setting Up the Recorderp.46Setting the Clockp.48Probe Detectp.48Setting the IP Addressp.49Adding a Measurement Descriptionp.49Configure Nominal Power Valuesp.50Using Phase Swapp.50Setting the Volts and Current Ratiop.51Assign Recorder Name and Passwordp.51Setting the Snapshot Period - Periodic Waveform Capture Settingp.52Change Display Language and Phase Identifierp.53Working with Annotationsp.53Using Image or Voice Annotationsp.55Inserting Image or Voice Annotationsp.55Turn off the Recorderp.56Cleaning and Maintenancep.56Regulatory Information for Wireless Communicationp.57Specifications for the System: Recorder and Power Analyze Softwarep.58General Specificationsp.58Input Specificationsp.58Synchronization and Samplingp.58Voltage and Current Measurementsp.58Voltage and Current Measurement Accuracyp.59Transient Voltage (Impulse)p.59Dip (Sag) and Swell Measurementsp.59Power and PF Measurementp.59External Interface Specificationsp.60Environmental and Safety Specificationsp.60
Page 1 of the Fluke 1750 user manual — 1750 Power Recorder
Page 1 · click to enlarge
Page 1

1750 Power Recorder

October 2006 Rev.3, 11/10 © 2006-2010 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies.

Page 2 of the Fluke 1750 user manual — LIMITED WARRANTY AND LIMITATION OF LIABILITY
Page 2 · click to enlarge
Page 2

LIMITED WARRANTY AND LIMITATION OF LIABILITY

Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year and begins on the date of shipment. Parts, product repairs, and services are warranted for 90 days. This warranty extends only to the original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries, or to any product which, in Fluke's opinion, has been misused, altered, neglected, contaminated, or damaged by accident or abnormal conditions of operation or handling. Fluke warrants that software will operate substantially in accordance with its functional specifications for 90 days and that it has been properly recorded on non-defective media. Fluke does not warrant that software will be error free or operate without interruption.

Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price. Fluke reserves the right to invoice Buyer for importation costs of repair/replacement parts when product purchased in one country is submitted for repair in another country.

Fluke's warranty obligation is limited, at Fluke's option, to refund of the purchase price, free of charge repair, or replacement of a defective product which is returned to a Fluke authorized service center within the warranty period.

To obtain warranty service, contact your nearest Fluke authorized service center to obtain return authorization information, then send the product to that service center, with a description of the difficulty, postage and insurance prepaid (FOB Destination). Fluke assumes no risk for damage in transit. Following warranty repair, the product will be returned to Buyer, transportation prepaid (FOB Destination). If Fluke determines that failure was caused by neglect, misuse, contamination, alteration, accident, or abnormal condition of operation or handling, including overvoltage failures caused by use outside the product’s specified rating, or normal wear and tear of mechanical components, Fluke will provide an estimate of repair costs and obtain authorization before commencing the work. Following repair, the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportation charges (FOB Shipping Point).

THIS WARRANTY IS BUYER'S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY.

Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a court or other decision-maker of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.

Fluke Corporation P.O. Box 9090 Everett, WA 98206-9090 U.S.A.

Fluke Europe B.V. P.O. Box 1186 5602 BD Eindhoven The Netherlands

To register your product online, visit register.fluke.com

11/99

Page 3 of the Fluke 1750 user manual — Table of Contents
Page 3 · Contents
Contents

Contents

1750 Operators Manualp.1LIMITED WARRANTY AND LIMITATION OF LIABILITYp.2Table of Contentsp.3List of Tablesp.5List of Figuresp.7Introductionp.9Contacting Flukep.10Safety Informationp.11Accessoriesp.12Featuresp.13Applying the Front Panel Decalp.15Charging the PDA Batteryp.17Installing the Softwarep.17Installing Power View on the PDAp.17Language Selectionp.18PC System Requirement for Power Analyzep.18Installing Power Analyzep.18Installing the Power Recorder at a Facilityp.19Work Flowp.19Installing the Recorderp.20Connecting the Recorder to the Wiringp.20Verifying Your Connectionp.21Finishing UPp.21Managing Recorded Datap.22Power Type Diagramsp.22Communicating with the Recorder Using Power Viewp.33Navigating in Power View on the PDAp.34Power View Menusp.34Icons on the Menu Barp.35Menu Treep.36Home Screenp.37Downloading Data and Erasing Memoryp.39Viewing Live Datap.43Scope Screenp.43Meter Screenp.44Phasor Screenp.44Harmonics Screenp.45Trend Screenp.45Viewing Phasesp.46Setting Up the Recorderp.46Setting the Clockp.48Probe Detectp.48Setting the IP Addressp.49Adding a Measurement Descriptionp.49Configure Nominal Power Valuesp.50Using Phase Swapp.50Setting the Volts and Current Ratiop.51Assign Recorder Name and Passwordp.51Setting the Snapshot Period - Periodic Waveform Capture Settingp.52Change Display Language and Phase Identifierp.53Working with Annotationsp.53Using Image or Voice Annotationsp.55Inserting Image or Voice Annotationsp.55Turn off the Recorderp.56Cleaning and Maintenancep.56Regulatory Information for Wireless Communicationp.57Specifications for the System: Recorder and Power Analyze Softwarep.58General Specificationsp.58Input Specificationsp.58Synchronization and Samplingp.58Voltage and Current Measurementsp.58Voltage and Current Measurement Accuracyp.59Transient Voltage (Impulse)p.59Dip (Sag) and Swell Measurementsp.59Power and PF Measurementp.59External Interface Specificationsp.60Environmental and Safety Specificationsp.60
Page 5 of the Fluke 1750 user manual — List of Tables
Page 5 · click to enlarge
Page 5

List of Tables

1. Symbols 4 2. Standard Accessories 4 3. Input Terminals and Controls 6

Table Title Page

Page 6 of the Fluke 1750 user manual — List of Tables
Page 6 · click to enlarge
Page 6
Page 6 of the Fluke 1750 user manual — List of Tables
Page 7 of the Fluke 1750 user manual — List of Figures
Page 7 · click to enlarge
Page 7

List of Figures

1. Fluke 1750 Power Recorder 5 2. Applying the Front Panel Decal 7 3. Connecting the Supplemental Ground Terminal 8 4. One Phase Plus Neutral 15 5. One-Phase IT No Neutral 16 6. One-Phase Split Phase 17 7. Three-Phase Wye 18 8. Three-Phase Delta 19 9. Three-Phase IT 20 10. Three-Phase High Leg 21 11. Three-Phase Open Leg 22 12. 2-Element Delta 23 13. 2 ½-Element Wye 24 14. Connecting to the Recorder 25 15. Power View Menu Tree 28

Figure Title Page

Page 8 of the Fluke 1750 user manual — List of Figures
Page 8 · click to enlarge
Page 8
Page 8 of the Fluke 1750 user manual — List of Figures
Page 9 of the Fluke 1750 user manual — Introduction
Page 9 · click to enlarge
Page 9

1750 Power Recorder

Introduction

The Fluke 1750 Power Recorder is a comprehensive yet easy to use system for power quality investigations. The 1750 Power Recorder, referred to hereafter as simply “the Recorder or the Product,” consists of a power recorder instrument, a wireless handheld Personal Digital Assistant (PDA) for control and setup, and a powerful yet easy to use PC application, Power Analyze. Four 400 A current probes are provided as standard equipment (not included in Basic version). A wide range of flexible and clamp-on current probes are available from Fluke.

Key features of the 1750 Power Recorder are:

  • No PC needed for setup

Using the included wireless-enabled Personal Digital Assistant (PDA) as a controller, you do not need a PC to configure the Recorder, troubleshoot connections, preview data, or download the recorded data. Downloaded data goes onto the SD memory card plugged into the 1750, not directly to the PDA. The range of the PDA with the wireless link is about 5 meters (16 feet) from the Recorder. This range is dependent on the electrical interference in the location.

  • Simplified test lead connections to the power network

Simply connect a voltage probe to a conductor on each phase that you want to record. On the current inputs, the Recorder automatically identifies what type of probe is connected. The Recorder then configures its measurement system appropriately for the model of current probe you are using.

  • Connection diagrams for the supported power network configurations (delta, Wye, and others) appear on the handheld PDA controller. Once you make connections, you can view live readings and a phasor diagram on the PDA to confirm the connections. If a phase channel is wired incorrectly, you can swap the phase to another channel by changing an internal Recorder setting using the PDA, and then reconfirm correct readings.
  • Internal Uninterruptible Power Supply (UPS)

An internal NiMH (Nickel-Metal-Hydride) battery pack and charging system maintain data capture continuity through power interruptions of approximately 5 minutes or less.

Page 10 of the Fluke 1750 user manual — Contacting Fluke
Page 10 · click to enlarge
Page 10
  • Automatic disturbance capture

The Recorder uses an automatic, self-learning threshold routine, which means you do not have to set any event threshold limits before you start recording. You detect and display power quality events (disturbances) using the Power Analyze software on a PC after the event has been captured. You define thresholds for events after data is stored on the PC. This is called the "event detector" in the software. It does not make permanent changes to the data file, so you can experiment with different thresholds after the fact without worrying about losing data.

With these features, there is little else required to set up and start recording the data to conduct a power quality survey. Should you want to extensively analyze collected data, the procedures described later in this manual review the transfer of data from the Recorder to the PDA, and finally to a PC. Data may then be analyzed using Power Analyze on a PC. Extensive online help is provided in the Power Analyze software on the product CD shipped with the Recorder.

Contacting Fluke

To contact Fluke, call one of the following telephone numbers:

  • Technical Support USA: 1-800-44-FLUKE (1-800-443-5853)
  • Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853)
  • Canada: 1-800-36-FLUKE (1-800-363-5853)
  • Europe: +31 402-675-200
  • Japan: +81-3-3434-0181
  • Singapore: +65-738-5655
  • Anywhere in the world: +1-425-446-5500

Or, visit Fluke's website at www.fluke.com

To register your product, visit http://register.fluke.com

To view, print, or download the latest manual supplement, visit http://us.fluke.com/usen/support/manuals

To contact Technical Supportfpqsupport@ fluke.com or 888-257-9897
Page 11 of the Fluke 1750 user manual — Safety Information
Page 11 · click to enlarge
Page 11

Power Recorder Safety Information

Safety Information

See the Specifications section for a full agency compliance list. Table 1 lists symbols that appear on the Recorder and in this manual.

XW Warning To prevent possible electrical shock, fire, or personal injury:

  • Carefully read all instructions.
  • Do not work alone.
  • Do not use the Product around explosive gas, vapor, or in damp or wet environments.
  • Use only insulated current probes, test leads, and adapters as supplied with the Product.
  • Examine the case before you use the Product. Look for cracks or missing plastic. Carefully look at the insulation around the terminals.
  • Remove all probes, test leads, and accessories that are not necessary for the measurement.
  • Make sure the ground conductor in the mains power cord is connected to a protective earth ground. Disruption of the protective earth could put voltage on the chassis that could cause death.
  • Only use probes, test leads, and accessories that have the same measurement category, voltage, and amperage ratings as the product.
  • Do not use exposed metal BNC or banana plug connectors.
  • Do not put metal objects into connectors.
  • Do not operate the Product with covers removed or the case open. Hazardous voltage exposure is possible.
  • There are no user-replaceable parts in the Product. Have an approved technician repair the Product.
  • Use the Product only as specified, or the protection supplied by the Product can be compromised.
Page 12 of the Fluke 1750 user manual — Accessories
Page 12 · click to enlarge
Page 12

Table 1. Symbols

SymbolMeaningSymbolMeaning XHazardous voltage. Risk of electrical shock.WRisk of danger. Important information. See manual. PConforms to requirements of European Union and European Free Trade Association (EFTA).~Do not dispose of this product as unsorted municipal waste. Contact Fluke or a qualified recycler for disposal. )Canadian Standards Association. [Note: Canadian and US.]JProtective conductor terminal. CAT IIICAT III equipment is designed to protect against transients in equipment in fixed-equipment installations, such as distribution panels, feeders and short branch circuits, and lighting systems in large buildings.CAT IVCAT IV equipment is designed to protect against transients from the primary supply level, such as an electricity meter or an overhead or underground utility service.

Accessories

Table 2 describes the standard accessories that ship with the Recorder.

Table 2. Standard Accessories

Fluke 1750 — Accessories diagram (page 12)
Fig. 1 · Accessories
DescriptionPart Number Ethernet cable, 3 meters, yellow2402854 Colored plastic clips for test leads (32 clips, 8 colors, 4 clips each color)2157607 Model TLS430, test lead set including cable and clips2157713 Secure Digital (SD) Memory Card for downloading dataCall for details Model 3140R, 400 A Clamps (4-not included with Basic version)2277216 AC power cord, 3 meters2441360 Personal Digital Assistant (PDA) with USB cable and accessories2386780 CD-ROM Manuals and Software2386771 1750 Getting Started Guide2386767 Sheet of Front Panel Decals2436261 Sets of international ac power plug adapters for the PDA charger2583479 International adapters for the Recorder power cord2441372
Page 13 of the Fluke 1750 user manual — Features
Page 13 · click to enlarge
Page 13

You can also order the following optional accessories. For more information about accessories, see Contacting Fluke earlier in this manual. An updated list is always available on www.fluke.com

  • Soft Carry Bag
  • Hard Transit Case
  • Security Cable
  • Replacement Ethernet Cable Kit
  • Various Current Probes
  • Replacement Voltage Lead Set
  • Clamp-on Current Transformers (3005R)
  • Flexible Current Transformers (3110, 3210, 3310, 3312)
  • Interface and Extension Cables 3570 CT Cable

Features

The Recorder front panel is illustrated in Figure 1 and described in Table 3.

Fluke 1750 — Features diagram (page 13)
Fig. 2 · Features

azd09f.eps Figure 1. Fluke 1750 Power Recorder

WCaution Connecting the supplemental ground terminal and the line cord safety ground to different ground potentials creates a ground loop that can damage the Recorder.

Power Recorder Features

Page 14 of the Fluke 1750 user manual — Features
Page 14 · click to enlarge
Page 14

Table 3. Input Terminals and Controls

A Standard 120/240 V 50/60 Hz power input for Recorder power

Steady Voltage is present, not overloaded Blinking Voltage overload (overload threshold 1100 V) C Five voltage measurement inputs

D Removable Secure Digital (SD) memory card for transfer of large quantities of data. You can remove the card from the Recorder, insert in the PDA and transfer data to the PC via the “synchronize data” feature of the PDA. A USB SD card reader may also be used.

E SD status indicator

Steady green SD card inserted Blinking Busy - do not remove the card

F Wireless Controller Status Indicator

Steady Blue Wireless controller enabled but not communicating Blinking Wireless controller communicating Off Inactive

G Ethernet port. TCP/IP via Ethernet is used to connect the Recorder to a PC for downloading recorded data, and for configuring the Recorder when not using the PDA. This is the recommended method of data transfer due to its speed and reliability.

Blinking Network is busy

Off No link Steady Link Present Blinking Communicating with PC

J Current Measurement Terminals (5)

Steady Current probe in normal range Blinking > 110% of probe range, use a higher range probe Off Current insufficient to measure

Steady green AC input power normal, instrument is not booted or locked Green/Orange blinking Recording Blinking red Recorder running on UPS

Connecting the supplemental ground terminal and the line cord safety ground to different ground potentials creates a ground loop that can damage the Recorder. To avoid this, use the supplemental ground terminal only when no protective earth ground connection exists through the line power cord. If there is any chance that a safety ground connection does exist through the line power cord connection, make sure to connect the supplemental ground terminal only to the ground system used by the ac receptacle that powers the Recorder. See Figure 3.

Show 7 more labels / short text on this page
Number Description · B Voltage indicator · H Busy Indicator · I Link indicator · K Current indicator · L Power LED · M Ground Terminal
Page 15 of the Fluke 1750 user manual — Applying the Front Panel Decal
Page 15 · click to enlarge
Page 15

Power Recorder Applying the Front Panel Decal

Applying the Front Panel Decal

Self-adhesive front panel decals are supplied with the 1750. The supplied decals correspond to the wiring color codes used in the USA, Europe and UK (new), Europe (old), UK, Canada, and China. Apply the decal appropriate for your local wiring codes around the current and voltage inputs as shown in Figure 2.

Fluke 1750 — Applying the Front Panel Decal diagram (page 15)
Fig. 3 · Applying the Front Panel Decal

azd17f.eps Figure 2. Applying the Front Panel Decal

Page 16 of the Fluke 1750 user manual — Applying the Front Panel Decal
Page 16 · click to enlarge
Page 16

WCaution Connecting the supplemental ground terminal and the line cord safety ground to different ground potentials creates a ground loop that can damage the Recorder.

Fluke 1750 — Applying the Front Panel Decal diagram (page 16)
Fig. 4 · Applying the Front Panel Decal

azd11f.eps Figure 3. Connecting the Supplemental Ground Terminal

Page 17 of the Fluke 1750 user manual — Installing the Software
Page 17 · click to enlarge
Page 17

Power Recorder Charging the PDA Battery

Charging the PDA Battery

Your PDA is shipped with a discharged or partially charged, rechargeable battery. You should fully charge the battery before setting up the device, and recharge it regularly. Refer to your PDA user’s guide for detailed information about the battery.

WCaution Do not leave important data on the PDA. If the battery discharges completely, you will lose all data not stored in File Store (ROM). Discharging the battery completely is similar to performing a hard reset. Refer to the PDA Users Manual for more information.

Installing the Software

This section describes how to install Fluke Power View software on the PDA and Fluke Power Analyze software on your PC. Keep your product CD in case you need to reinstall this software in the future. You should periodically check the Fluke website: www.fluke.com to see if any 1750 firmware or application software updates are available for download. Fluke encourages you to register the Recorder. After your Recorder has been registered, you will receive email notifications of firmware and software update availability.

Installing Power View on the PDA

1. Insert the CD that ships with your PDA. Follow the on-screen instructions to install the ActiveSync software for Windows XP (or Mobile Device Center for Windows Vista and Windows 7) that allows your PC to communicate with the PDA.
2. Connect the PDA to your PC with the USB lead and establish communication between the PDA and your PC. Keep your product CD in case you need to reinstall this software in the future.
3. Insert the CD that ships with your Recorder in the CD-ROM drive on your PC.
4. A window appears that lists options on the CD. If the install software window does not automatically appear, do the following:

a. Use Windows Explorer to locate the sw/FPV_Vx.yy.exe file on the CD. Replace x.yy

with the current version number shipped on the CD.

b. Double-click FPV_Vx.yy_Install.exe to launch the installation program.

5. Select a language preference for the installation.
6. Follow the on-screen prompts.

Power View is installed in the Programs folder on the PDA.

7. To start Power View, select Start, and then Programs on the PDA.
8. Tap the Power View icon (

) to start using the application.

Page 18 of the Fluke 1750 user manual — Language Selection
Page 18 · click to enlarge
Page 18

Power View features a localized user interface for English, German, French, Spanish, Chinese, and Italian. When Power View operates for the first time on a new PDA or after a hard reset, it directly opens the language setup window. This window is also accessed by Menu>1750 Setup>Language.

The Language window also provides the option to choose between the phase identifiers A, B, C and L1, L2, L3.

Fluke 1750 — Language Selection diagram (page 18)
Fig. 5 · Language Selection

PC System Requirement for Power Analyze

  • Windows 2000, Windows XP, Windows Vista 32/64bit, Windows 7 32/64bit. Administrator privileges are required to install applications.
  • PC with 500 MHz or faster processor.
  • 256 MB RAM, 1 GB recommended (100 MB of available disk space, 10 GB recommended).
  • 1024 x 768 or higher resolution video.
  • Keyboard and mouse.

Before installing the software on your PC, make sure you have Internet Explorer version 6, or later, installed. Other recommended software:

  • Adobe Acrobat Reader
  • Microsoft Word 2000 or higher
  • Microsoft Excel 2000 or higher

WCaution Do not remove the CD until after you have successfully installed the software, rebooted the machine, and opened the application.

1. Insert the CD into your CD-ROM drive.

Language Selection · Installing Power Analyze

Page 19 of the Fluke 1750 user manual — Installing the Power Recorder at a Facility
Page 19 · click to enlarge
Page 19

Power Recorder Installing the Power Recorder at a Facility

2. A window appears that lists options on the CD. If the window does not automatically appear:

a. Use Windows Explorer to locate the sw/FPV_Vx.yy.exe file on the CD.

Replace x.yy with the current version number shipped on the CD.

b. Double-click FPA_Vx.yy_Install.exe.

3. Select a language preference for the installation process. Power Analyze initially starts in this language. You can change the language from the menu Settings>Language .
4. Follow the installation instructions that appear. After software installation is complete, the Power Analyze icon

appears on your desktop.

5. Start Power Analyze on your PC by choosing Start > All Programs > Fluke > Power Analyze > Launch Power Analyze or double-click the Power Analyze icon

on your desktop.

6. After the installation is complete, please take a few minutes to complete your product registration. You can register quickly online (preferred) at http://register.fluke.com or you can print the form and fax it to the number provided on the form. Product registration is very important since it allows you to get free software updates and helps us provide you with the fastest and most efficient technical support.

Installing the Power Recorder at a Facility This section describes the steps you need to take to set up the Recorder at a facility and begin a recording session.

There are three distinct stages for a recording session.

  • Setup Setup, hookup, and verification of connections and measured signals. This is when you will want to clear the Recorder memory (not required, but best practice because data download when finished only contains your new data). This is a good time to change any recorder settings. Edit the measurement description now or after downloading the measurement data and insert annotations and/or an optional START mark into the recorded data stream. Annotations are messages or notes you insert into the data stream during a recording session.
  • Interim checkup, mid-recording session At this time, you may want to re-check all the live input signals, and download the data recorded since the beginning of the recording session. This does not interfere with ongoing recording in any way. Annotation capability is always available during the recording session.
  • End of recording session You can re-verify if desired, correctness of live data and then download data, power off the Recorder, and pack up for transport.

Work Flow

Page 20 of the Fluke 1750 user manual — Installing the Recorder
Page 20 · click to enlarge
Page 20
1. Position the Recorder within 2 m (6 ft) of the monitoring location.
2. The Recorder can be set on the floor or a table, or attached and secured to a pole or other mounting surface. The Recorder can be oriented vertically or horizontally.
3. Connect the Recorder to a 100 to 240 V ac power source, 50 or 60 Hz.
4. Plug the power cord into a properly-grounded wall outlet.

WCaution To avoid damage to the Recorder:

  • Ground the chassis as appropriate according to your power connection before you do anything else.
  • Connect the power cord to the Recorder using the appropriate supplied adapter.
  • Connect the power cord to a properly grounded wall outlet.
  • Be sure to plug the power cord into the Recorder panel BEFORE connecting it to an outlet.

After a minute, or two, all LEDs should flash ON, then OFF, and then each should turn ON and OFF in sequence. The LEDs will then remain on if you make secure test lead connections and there is sufficient voltage and current.

5. Turn on the PDA and tap the Power View icon

to launch Power View.

If a Recorder is within range and is not password protected, it should appear in the drop down menu just below the main title bar on the PDA. You may need to select Browse for recorder when first connecting. Select the recorder you want to browse to with the PDA stylus. Upon selection, the Home screen for live data appears.

If more than one Recorder is within range, these will appear in the drop down menu list.

6. If you have previously established password protection for the Recorder, you must type the correct password in the Password text box and click OK . (Password protection is established for a specific Recorder using either the Setup Password menu in Power View or the Power Analyze software).

Connecting the Recorder to the Wiring

After you have set up the Recorder you are ready to connect the voltage leads and current probe or flexi-probes to the power network being tested. Refer to the Power Type diagrams provided on the PDA or in this manual.

XWWarning To prevent possible electrical shock, Flexi-probes should not be applied or removed from hazardous live conductors.

In most cases, you should install the current probes first. Because they are clamped around wires, current probes are usually more secure than voltage probes. Examine the conductors you are about to connect to and determine if you should attach the current probes to the phase wires or busbars before you connect the voltage probes.

1. Select and attach the appropriate current probes to the Recorder.
2. Select and attach the appropriate voltage test leads and probes to the Recorder.

Installing the Recorder

Page 21 of the Fluke 1750 user manual — Verifying Your Connection
Page 21 · click to enlarge
Page 21

Power Recorder Installing the Power Recorder at a Facility

3. Connect the Recorder to the wiring by first connecting the measurement ground test lead to the distribution system ground, and then the neutral probe to neutral.
4. Connect the ground current probe to the distribution system ground and the neutral current probe to neutral, with the arrow on the current probe pointing towards the load (away from the source) in each case.
5. Connect the remaining current probes to the power network.
  • The arrow on the current probe should point toward the load.
  • Use the color-coding on the leads to help you connect the probes to the correct phases, neutral and ground.
6. Connect the remaining voltage probes.
7. Check the LED for each phase to make sure a connection is established.
  • When a phase LED is ON or lighted, you know that the connection is active and that voltage and current is within the normal range.
  • When a phase LED is BLINKING, you are connected but the voltage or current is exceeding the normal range or exceeding the range of the probe that you are using.
  • When a current LED is OFF or unlit, there is insufficient current present in the line.

Before you leave the Recorder to accumulate data at the facility, double-check the connections.

  • Make sure the current probes are secured and completely closed around the conductors and that the voltage and current LEDs are steady-on.
  • Use the PHASOR screen on the PDA or PC to verify that the voltage and current connections are correct. You can swap the connections using Power View if they are incorrect.

After making any corrections to the connections, recheck the PHASOR diagrams to confirm your connection. Make sure the nominal voltage, nominal line frequency, and power type (delta, wye, other) are correct before you leave the Recorder. It is a good practice to insert a Start Mark when you are finished with all settings so you can download data and ignore useless readings and events common at the beginning probe connecting activities. You can also use the “Erase Memory” function and measurement

Verifying Your Connection · Finishing UP

Page 22 of the Fluke 1750 user manual — Power Type Diagrams
Page 22 · click to enlarge
Page 22

descriptions such as company, name, and measurement purpose. This can be done at this point or after downloading the measurement data.

Recorded data resides in the Recorder on an internal non-physically accessible flash memory circuit. It is not transferred to an SD memory card in the Recorder’s SD memory slot until you request a download action using the PDA or an attached PC. Downloading to a PC running Fluke Power Analyze Software (included) using the Ethernet cable is recommended and the fastest way to retrieve data from the Recorder.

It is never technically necessary to erase internal memory. When the Recorder has new data to record, it does so automatically, overwriting the oldest data (circular memory). You can erase internal memory for security reasons, or if you want to simplify what you will be seeing in the download data screen. The start and end times will represent one recording session when old data is erased.

Power Type Diagrams

The wiring illustrations included cover the standard power configurations that are selectable in Power View or Power Analyze software. The diagrams are provided as an aid in making the correct test lead connections. Power type diagrams are also included on your PDA and can be referenced when you are setting up nominal power on the Recorder.

Available power type settings are:

  • One-Phase Plus Neutral (Figure 4)
  • One-Phase IT No Neutral (Figure 5)
  • One-Phase-Split Phase (Figure 6)
  • Three-Phase Wye (Figure 7)
  • Three-Phase Delta (Figure 8)
  • Three-Phase IT (Figure 9)
  • Three-Phase High Leg (Figure 10)
  • Three-Phase Open Leg (Figure 11)
  • 2-Element Delta (Figure 12)
  • 2 ½-Element Wye (Figure 13)

Managing Recorded Data

Page 23 of the Fluke 1750 user manual — Managing Recorded Data
Page 23 · click to enlarge
Page 23

Power Recorder Power Type Diagrams

Fluke 1750 — Managing Recorded Data diagram (page 23)
Fig. 6 · Managing Recorded Data

azd02f.eps Figure 4. One Phase Plus Neutral

Example: Branch circuit at an outlet.

Page 24 of the Fluke 1750 user manual — Managing Recorded Data
Page 24 · click to enlarge
Page 24
Fluke 1750 — Managing Recorded Data diagram (page 24)
Fig. 7 · Managing Recorded Data

azd14f.eps Figure 5. One-Phase IT No Neutral

Example: Used in Norway and in some hospitals. This would be the connection at a branch circuit.

Page 25 of the Fluke 1750 user manual — Managing Recorded Data
Page 25 · click to enlarge
Page 25

Power Recorder Power Type Diagrams

Fluke 1750 — Managing Recorded Data diagram (page 25)
Fig. 8 · Managing Recorded Data

azd03f.eps Figure 6. One-Phase Split Phase

Example: A North American residential installation at the service entrance.

Page 26 of the Fluke 1750 user manual — Managing Recorded Data
Page 26 · click to enlarge
Page 26
Fluke 1750 — Managing Recorded Data diagram (page 26)
Fig. 9 · Managing Recorded Data

azd04f.eps Figure 7. Three-Phase Wye

Example: Also called “Star” or four-wire connection. Typical commercial building power.

Page 27 of the Fluke 1750 user manual — Managing Recorded Data
Page 27 · click to enlarge
Page 27

Power Recorder Power Type Diagrams

Fluke 1750 — Managing Recorded Data diagram (page 27)
Fig. 10 · Managing Recorded Data

azd05f.eps Figure 8. Three-Phase Delta

Example: Often found in industrial settings where electric motors are used.

Page 28 of the Fluke 1750 user manual — Managing Recorded Data
Page 28 · click to enlarge
Page 28
Fluke 1750 — Managing Recorded Data diagram (page 28)
Fig. 11 · Managing Recorded Data

azd12f.eps Figure 9. Three-Phase IT

Example: Industrial power in countries that use the IT (Isolated Terra) system, such as Norway.

Page 29 of the Fluke 1750 user manual — Managing Recorded Data
Page 29 · click to enlarge
Page 29

Power Recorder Power Type Diagrams

Fluke 1750 — Managing Recorded Data diagram (page 29)
Fig. 12 · Managing Recorded Data

azd08f.eps Figure 10. Three-Phase High Leg

Example: A way to get a 120 V single phase feed by tapping across part of a leg in a delta power system

Page 30 of the Fluke 1750 user manual — Managing Recorded Data
Page 30 · click to enlarge
Page 30
Fluke 1750 — Managing Recorded Data diagram (page 30)
Fig. 13 · Managing Recorded Data

azd18f.eps Figure 11. Three-Phase Open Leg

Example: A variant of power transformer winding type.

Page 31 of the Fluke 1750 user manual — Managing Recorded Data
Page 31 · click to enlarge
Page 31

Power Recorder Power Type Diagrams

Fluke 1750 — Managing Recorded Data diagram (page 31)
Fig. 14 · Managing Recorded Data

azd13f.eps Figure 12. 2-Element Delta

Example: Blondel or Aron connection, used to measure power like a two-element revenue power meter.

Page 32 of the Fluke 1750 user manual — Managing Recorded Data
Page 32 · click to enlarge
Page 32
Fluke 1750 — Managing Recorded Data diagram (page 32)
Fig. 15 · Managing Recorded Data

azd07f.eps Figure 13. 2 ½-Element Wye

Example: Blondel or Aron system for 4-wire (Wye) power secondaries.

Page 33 of the Fluke 1750 user manual — Communicating with the Recorder Using Power View
Page 33 · click to enlarge
Page 33

Power Recorder Communicating with the Recorder Using Power View

Communicating with the Recorder Using Power View

The PDA serves as your control panel for the Recorder. The Power View software installed on the PDA is designed to function as a remote front panel for the Recorder, since the Recorder itself is a “black box” with no integral display or external controls other than status LEDs. With Power View installed on the PDA, you are relieved of the burden of having to carry a laptop and Ethernet cable to a remote site when setting up the Recorder, checking its progress, or finishing a monitoring session.

You can alternatively connect to a Recorder from a PC directly using an Ethernet cable or remotely over an IP network. You can then use Fluke Power Analyze to set up the Recorder, view power measurements wherever the Recorder is installed, or download data from the Recorder.

Power View communicates with the PDA using a wireless controller radio interface. There is no provision for a hardwired PDA interface to the Ethernet port on the Recorder. After data is recorded and downloaded to a PC, you can analyze the recorded data using the Power Analyze software.

Fluke 1750 — Communicating with the Recorder Using Power View diagram (page 33)
Fig. 16 · Communicating with the Recorder Using Power View

azd01f.emf Figure 14. Connecting to the Recorder

Page 34 of the Fluke 1750 user manual — Navigating in Power View on the PDA
Page 34 · click to enlarge
Page 34

Navigating in Power View on the PDA

The Menu Bar on the bottom of the Power View screen allows you to quickly and easily configure a Recorder, view live data, and transfer data to the Secure Data (SD) card for viewing and analysis on your desktop computer using the companion Power Analyze software.

Brings up a list of the live detail views: Scope, Meter, Phasor, Harmonics, and Trend. There are configurable detail views of each type of screen available by tapping the sector of interest or the Live menu. Once a detail view is selected, navigation to all available detail views is accomplished using the Live menu.

Brings up the Data memory management screens which are for download to the SD card inserted in the Recorder and for erasing Recorder internal memory and erasing SD card files.

Brings up the setup screen that allows you to configure the Recorder and set recording period information. Using this function, you can tap the appropriate icon to set the following:

  • Clock
  • Probe Detect
  • IP Address
  • Measurement Description
  • Nominal Power
  • Phase Swap
  • Volts/Current Ratio
  • Recorder Name/Password
  • Snapshot Period
  • Language
Show 4 more labels / short text on this page
Power View Menus · Live · Menu · 1750 Setup
Page 35 of the Fluke 1750 user manual — Icons on the Menu Bar
Page 35 · click to enlarge
Page 35

Power Recorder Navigating in Power View on the PDA

Icons on the Menu Bar

are used to toggle between voltage and current wave forms. Tap either V or A and then Live>Home to see each parameter’s detailed data. The PHASOR screen always shows both V & A (volts and amps).

on the detail views, as voltage and current data can be viewed simultaneously. In all the detail live view screens, a line selection menu is available using the

button at the bottom. This is a checkbox that is used to select the voltage and current channels you want to display.

The Snapshot icon triggers the Recorder to capture high-resolution waveform data and save it in the recorded data stream. This waveform capture is identical to the snapshots taken automatically when you set the “snapshot period” in the 1750 Setup menu.

This icon brings up the Annotate screen. Use the annotate menu to enter text or to insert a file name and load it into the recorded data stream in the Recorder. Annotations can be used to make notes, or to insert a file name of an image or audio file. You can enter multiple start and stop annotations. START and stop annotations are not paired and are not restricted to being used as literally “start” and “stop”. Annotations are often done at the start of a power recording session.

and · is replaced with

Page 36 of the Fluke 1750 user manual — Menu Tree
Page 36 · click to enlarge
Page 36

Menu Tree

The following figure provides an overview of the Power View menu structure that will be helpful in navigating through the Power View application.

Fluke 1750 — Menu Tree diagram (page 36)
Fig. 17 · Menu Tree

azd15f.eps Figure 15. Power View Menu Tree

Page 37 of the Fluke 1750 user manual — Home Screen
Page 37 · click to enlarge
Page 37

Power Recorder Home Screen

Home Screen

The Home screen is the top-level screen, accessed when you start Power View or when you tap Live and then Home on the Menu Bar.

The Home screen contains a 1.5 cycle waveform screen, a digital meter screen, and a PHASOR screen. The parameters shown in this screen are not configurable. All phases are shown simultaneously, either all voltage or all current.

Fluke 1750 — Home Screen diagram (page 37)
Fig. 18 · Home Screen

Azd111.bmp If no Recorders are within range, the following screen appears:

Fluke 1750 — Home Screen diagram (page 37)
Fig. 19 · Home Screen
Page 38 of the Fluke 1750 user manual — Home Screen
Page 38 · click to enlarge
Page 38

If more than one Recorder is within range, the Recorder selection screen appears. An asterisk preceding the Recorder name means that unit is password protected. If your recorder is not listed, but in range, press "Search New Devices" to update the list.

Fluke 1750 — Home Screen diagram (page 38)
Fig. 20 · Home Screen

If the selected Recorder is not password protected, the Home screen appears. If password protection is set, the Password screen appears instead.

Fluke 1750 — Home Screen diagram (page 38)
Fig. 21 · Home Screen
Page 39 of the Fluke 1750 user manual — Downloading Data and Erasing Memory
Page 39 · click to enlarge
Page 39

Power Recorder Downloading Data and Erasing Memory

If the password is incorrect the password text field is cleared and then you can re-enter the password.When the password is accepted, the Home screen appears.

Downloading Data and Erasing Memory

Use the Data menu options to manage the Recorder memory and download recorded data to the SD card plugged into the Recorder's (not the PDA) SD card slot. You are downloading recorded data from the flash memory to an SD card in the Recorder. When you want to use that data, you must move the card to the PDA. You can later copy the data to your PC for review using the Power Analyze software. Tap 1750 Internal Memory in the Data menu to view the Download screen.

Fluke 1750 — Downloading Data and Erasing Memory diagram (page 39)
Fig. 22 · Downloading Data and Erasing Memory

The Download screen shows a three-phase rms V strip chart with date and timescale shown. This represents the data in the Recorder internal memory. The default settings of the start and stop times is everything in memory. To download ALL, tap the Write to SD button. If you want to trim off data that contains partial information because probes were being connected, select the START mark annotation (if one was entered by the user) as the start point.

The gray cursor lines right and left with triangular arrowheads at the bottom of the screen show the start and stop times currently selected for download. Green and red flags are START and STOP mark annotations inserted during the recording session. Yellow flags are file name annotations inserted during the recording session, or setup parameter changes that change the data significantly, such as nominal voltage or power type.

Page 40 of the Fluke 1750 user manual — Downloading Data and Erasing Memory
Page 40 · click to enlarge
Page 40
Fluke 1750 — Downloading Data and Erasing Memory diagram (page 40)
Fig. 23 · Downloading Data and Erasing Memory

If the SD storage card already contains data files, it may not have room for the new download. Power View alerts you to this condition and asks if you would like to erase the SD card before downloading.

Fluke 1750 — Downloading Data and Erasing Memory diagram (page 40)
Fig. 24 · Downloading Data and Erasing Memory

Azd107s.bmp

You can also access the SD card to delete files by selecting Tools>1750 SD Memory from the Data menu option.

Page 41 of the Fluke 1750 user manual — Downloading Data and Erasing Memory
Page 41 · click to enlarge
Page 41

Power Recorder Downloading Data and Erasing Memory

You can delete files as necessary, and when satisfied that there is room on the card, proceed to write the selected internal Recorder memory data to the SD card.

Fluke 1750 — Downloading Data and Erasing Memory diagram (page 41)
Fig. 25 · Downloading Data and Erasing Memory

Azd108s.bmp

When there is room on the card, a message confirming the download will appear.

Fluke 1750 — Downloading Data and Erasing Memory diagram (page 41)
Fig. 26 · Downloading Data and Erasing Memory

You can also tap the Erase 1750 Memory from the Data menu option which is available under Menu, and tap the Erase All button to erase all power quality data stored in the Recorder. Even then, if you want, you can leave old data in Recorder memory. It will be overwritten automatically when you start your next Recorder session, so it never has an adverse effect on recorded data quality or quantity.

Page 42 of the Fluke 1750 user manual — Downloading Data and Erasing Memory
Page 42 · click to enlarge
Page 42
Fluke 1750 — Downloading Data and Erasing Memory diagram (page 42)
Fig. 27 · Downloading Data and Erasing Memory

Azd109s.bmp

When you choose to Erase All, the following popup caution “Are you sure?” appears on the screen.

Fluke 1750 — Downloading Data and Erasing Memory diagram (page 42)
Fig. 28 · Downloading Data and Erasing Memory
Page 43 of the Fluke 1750 user manual — Viewing Live Data
Page 43 · click to enlarge
Page 43

Power Recorder Viewing Live Data

Viewing Live Data

The Live menu is used to view live data on the circuits on the circuits for example, Scope, Meter, Phasor, or Harmonics.

Fluke 1750 — Viewing Live Data diagram (page 43)
Fig. 29 · Viewing Live Data

The Scope screen shows 1.5 cycle waveform data for the volts and amp lines selected in the check boxes on Phase View screen. Tap the Phase icon (

) to display the Phase View screen.

Fluke 1750 — Viewing Live Data diagram (page 43)
Fig. 30 · Viewing Live Data

Scope Screen

Page 44 of the Fluke 1750 user manual — Meter Screen
Page 44 · click to enlarge
Page 44

The Meter screen shows numeric readings for all selected phases.

Fluke 1750 — Meter Screen diagram (page 44)
Fig. 31 · Meter Screen

The Phasor screen shows voltage (long arrows with solid arrowheads), current amplitudes, and phase angles. This screen is used to verify proper test lead and current probe connections. You can use the V and A arrows on the upper right of the screen to change the scaling of voltage and current on the PHASOR diagram.

You can easily correct an erroneous Recorder connection by going directly to Menu\1750 Setup\Phase Swap (

) and using the Recorder to internally invert a CT, swap a voltage or CT phase, or modify a scaling parameter, then return to the screen to verify that the correction has been made. There is no need to touch the cables. See “Phase Swap” later in the manual for additional information.

Fluke 1750 — Meter Screen diagram (page 44)
Fig. 32 · Meter Screen

Meter Screen · Phasor Screen

Page 45 of the Fluke 1750 user manual — Harmonics Screen
Page 45 · click to enlarge
Page 45

Power Recorder Viewing Live Data

The Harmonics screen shows the magnitudes of the selected phases of V and I. The scale is % of nominal.

The slider control displays harmonics of a higher order up to the 50th. The V and A scaling arrows are on the screen in the upper right.

If the power type is delta, V phases are AB, BC, and CA (L1L2, L2L3, L3L1 is available as a localized European labeling scheme). If the power type is Wye, V phases are AN, BN, CN, and NG. (N = neutral, G = ground).

Fluke 1750 — Harmonics Screen diagram (page 45)
Fig. 33 · Harmonics Screen

The Trend screen shows live data being recorded as a strip chart display of the selected phases of voltage or current. This moment in time is visible on the right side of the graph.

The traces fill the screen in 10 minutes, and continue FIFO (first-in, first-out). This feature is intended to paint a picture of the short-term stability of the power system. The V and A scaling arrows on the upper right work like the scaling arrows on the PHASOR screen.

Fluke 1750 — Harmonics Screen diagram (page 45)
Fig. 34 · Harmonics Screen

Harmonics Screen · Trend Screen

Page 46 of the Fluke 1750 user manual — Setting Up the Recorder
Page 46 · click to enlarge
Page 46

In all the detail Live View screen, a line selection menu is available using the

button at the bottom. The Phase Selection screen reveals check boxes for selecting the voltage and current channels you want to display.

Fluke 1750 — Viewing Phases diagram (page 46)
Fig. 35 · Viewing Phases

Setting Up the Recorder

To adjust the settings of the Recorder, tap Menu>1750 Setup.

Fluke 1750 — Setting Up the Recorder diagram (page 46)
Fig. 36 · Setting Up the Recorder

Viewing Phases

Page 47 of the Fluke 1750 user manual — Viewing Phases
Page 47 · click to enlarge
Page 47

Power Recorder Setting Up the Recorder

The Setup screen shows setup options.

Fluke 1750 — Viewing Phases diagram (page 47)
Fig. 37 · Viewing Phases

From the Setup screen, you can adjust either the recording period information or the Recorder settings. Tap an icon on the Setup screen to view or configure:

  • Clock
  • Probe Detect
  • IP Address
  • Measurement Description
  • Nominal Power
  • Phase Swap
  • Volts/Current Ratio
  • Recorder Name/Password
  • Snapshot Period
  • Language
Page 48 of the Fluke 1750 user manual — Setting the Clock
Page 48 · click to enlarge
Page 48

Use this feature to set the PDA clock or clock to synchronize the Recorder with the PDA clock.

Fluke 1750 — Setting the Clock diagram (page 48)
Fig. 38 · Setting the Clock

Azd128.bpm

This screen shows the types of probes attached to the Recorder current inputs. Current probes are automatically detected.

Fluke 1750 — Setting the Clock diagram (page 48)
Fig. 39 · Setting the Clock

Setting the Clock · Probe Detect

Page 49 of the Fluke 1750 user manual — Setting the IP Address
Page 49 · click to enlarge
Page 49

Power Recorder Setting Up the Recorder

Setting the IP Address

Select Obtain Address Automatically unless your company's IT department has specific requirements that require assigned IP addresses.

Fluke 1750 — Setting the IP Address diagram (page 49)
Fig. 40 · Setting the IP Address

Adding a Measurement Description

Click on the arrow and select one of the six available captions from the list box and enter the text in the text field below. The number of characters is limited to 50. Click on Insert to apply the changes.

You may change or delete the caption. The maximum possible length for caption text is 20 characters.

To apply changes, click on Apply.

Click OK to close the window and send the measurement description to the device.

Fluke 1750 — Setting the IP Address diagram (page 49)
Fig. 41 · Setting the IP Address
Page 50 of the Fluke 1750 user manual — Configure Nominal Power Values
Page 50 · click to enlarge
Page 50

Configure Nominal Power Values

This window contains the setup for Power Config, Nominal Voltage, and Frequency. It is important to correctly configure these values before the measurement as some calculations are based on these settings and cannot be recalculated after the measurement has been performed. Click on the 3-phase delta icon to display the wiring diagram of the selected topology. The list box of nominal voltage levels contains commonly used voltage levels. Different voltage levels may be entered if required.

Fluke 1750 — Configure Nominal Power Values diagram (page 50)
Fig. 42 · Configure Nominal Power Values
  • Rather than having to change the physical connections to the power system, use the Phase Swap feature to make a software-selectable correction without physically switching the cable. You can select from the following:
  • Swap any voltage input
  • Swap any current input
  • Invert any current input clamp

The screen shows check boxes to invert any current probe. Current probes are directional, and selecting “invert” has the effect of removing the probe and turning it around.

If the power type is delta, V phases are AB, BC, and CA (L1L2, L2L3, L3L1 is available as an optional setting). If the power type is Wye, V phases are AN, BN, CN, and NG. (N = neutral, G = ground).

Using Phase Swap

Page 51 of the Fluke 1750 user manual — Setting the Volts and Current Ratio
Page 51 · click to enlarge
Page 51

Power Recorder Setting Up the Recorder

Fluke 1750 — Setting the Volts and Current Ratio diagram (page 51)
Fig. 43 · Setting the Volts and Current Ratio

Setting the Volts and Current Ratio

Use this menu to add a ratio factor to the voltage or current input on each phase. Changing the first “1” in each case if there is a potential transformer in series with the voltage connection such as when you want to monitor a medium-voltage network using PTs and CTs to scale the voltage and current to a convenient level for measuring instruments.

The current ratio applies a ratio to the current input in order to show the current present on a medium-voltage network. This feature is intended to be used for sensing low levels of current to represent the much higher levels on the primary side at a substation or step-down transformer that has built-in metering current transformers.

Fluke 1750 — Setting the Volts and Current Ratio diagram (page 51)
Fig. 44 · Setting the Volts and Current Ratio

Assign Recorder Name and Password

Use this feature to assign a name identifier to the Recorder and to establish a password for the Recorder. This naming feature is designed to help you keep track of which Recorder is in use. The Recorder name is later associated with the data file that you download from the Recorder.

Page 52 of the Fluke 1750 user manual — Setting the Snapshot Period - Periodic Waveform Capture Setting
Page 52 · click to enlarge
Page 52
Fluke 1750 — Setting the Snapshot Period - Periodic Waveform Capture Setting diagram (page 52)
Fig. 45 · Setting the Snapshot Period - Periodic Waveform Capt…

Setting the Snapshot Period - Periodic Waveform Capture Setting

Use the Snapshot Period screen to program the Recorder to take automatic power conditions readings which are then stored in the recorded data stream. These snapshots contain a phasor diagram, waveforms of current and voltage, harmonics profiles, and other data.

  • None
  • 10 minutes
  • 30 minutes
  • 1 hour

on any of the View screens to take a snapshot. The message, “Snapshot Acquired” confirms that a snapshot was taken.

Fluke 1750 — Setting the Snapshot Period - Periodic Waveform Capture Setting diagram (page 52)
Fig. 46 · Setting the Snapshot Period - Periodic Waveform Capt…
Page 53 of the Fluke 1750 user manual — Change Display Language and Phase Identifier
Page 53 · click to enlarge
Page 53

Power Recorder Setting Up the Recorder

Fluke 1750 — Change Display Language and Phase Identifier diagram (page 53)
Fig. 47 · Change Display Language and Phase Identifier

Change Display Language and Phase Identifier

Power View features a localized user interface for English, German, French, Spanish, Chinese, and Italian. Select one of these languages and select your preference of phase identifier as A, B, C or L1, L2, L3. A restart of Power View is required to apply these changes.

Fluke 1750 — Change Display Language and Phase Identifier diagram (page 53)
Fig. 48 · Change Display Language and Phase Identifier

Annotations are optional markers that can be inserted by the Recorder operator to flag a particular point in the data when recording. They are usually inserted on using Power View with the wireless PDA.

Working with Annotations

Page 54 of the Fluke 1750 user manual — Change Display Language and Phase Identifier
Page 54 · click to enlarge
Page 54
Fluke 1750 — Change Display Language and Phase Identifier diagram (page 54)
Fig. 49 · Change Display Language and Phase Identifier

There are five types of annotations:

  • START annotation (green)

The START annotation can be used to signify where important data begins. For example, the operator might want to establish a start annotation after having finished all probe connections and instrument setup, to indicate that any preceding data was preliminary.

  • STOP annotation (red)

The STOP annotation can be used to signify where important data ends.

  • File annotation (yellow)

File annotations contain the name of a picture or audio file relevant to the data. See “Using Image or Voice Annotations” later in this manual.

  • Text annotation

Text annotations contain a text note.

  • Power Configuration Change annotation

Power Configuration Change annotations are inserted by the system when a change is made to the Power Configuration in the Recorder. These markers only display in the Download dialog box.

To show or hide annotation markers:

  • In the Volts/Amps/Hz or Events view, check or uncheck the Annotations option on the view controls.

To open an annotation:

  • In the Volts/Amps/Hz view, Events view, or Download dialog box, click on the annotation.

If the annotation is associated with an audio (.WAV) or image (.jpg) file, the application associated with that file will start and open the file. If the file is not in the

Page 55 of the Fluke 1750 user manual — Using Image or Voice Annotations
Page 55 · click to enlarge
Page 55

Power Recorder Setting Up the Recorder

default directory My Documents\Fluke\PQ, a dialog box opens so that you can navigate to the file.

To see overlapping annotations:

Sometimes annotations are so close to each other that they actually overlap on the Volts/Amps/Hz reference graph. When this is the case, if you click on overlapping annotations, a menu opens and you can select either Open, to open the annotation now on top, or Select Next Overlapping Annotation, to bring the next one to the top.

  • To open the annotation now on top, choose Open.
  • To bring the next annotation to the top, choose Select Next Overlapping Annotation. To then open that annotation, click on it again and choose Open.

Using Image or Voice Annotations

While you are monitoring data with a Recorder, you may find it helpful to take a picture that relates to the power quality data you are recording, or to record verbal comments as you observe the conditions or review the data being measured. You can then associate the image or recording with the data being recorded by adding annotations that reference them.

The PDA includes a microphone for recording notes which are saved as .WAV files, and an optional camera accessory is available on the PDA which will capture images as .jpg files.

You can associate both images and audio files with the data being gathered in the Recorder by inserting annotations and specifying the name of a .WAV or .jpg file. After you download data from the Recorder, you can then move the image and audio files to the PC along with the recorded data (.odn) file.

Inserting Image or Voice Annotations

1. Set up the Recorder and begin monitoring.
2. Using your PDA-cam (not available in standard PDA supplied) or another camera, take a picture of what you want to capture.
3. Click the Annotate icon to display the Annotate screen.
4. Move the image or audio file to a location that will make it easier to Transfer it to a PC later, when you download and move the recorded data file to your PC. Place the
Page 56 of the Fluke 1750 user manual — Cleaning and Maintenance
Page 56 · click to enlarge
Page 56

file into the synchronization folder on your PDA, so that it will be automatically transferred into the sync folder on the PC the next time you dock the PDA.

Turn off the Recorder

The Recorder always has the UPS enabled to allow a 5-minute pass through time on a voltage interruption. When you pack the Recorder for transport, you have the option to shut down the Recorder and disable the UPS before unplugging the power cord.

To turn off the Recorder:

1. Select Menu > Tools > 1750 Shut down...
2. Confirm the warning message.
3. Wait until the Power LED is solid green.

The power cord can be safely unplugged.

Fluke 1750 — Turn off the Recorder diagram (page 56)
Fig. 50 · Turn off the Recorder

Cleaning and Maintenance

The Recorder contains no user serviceable parts. Contact an Authorized Fluke Service Center for repair. Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents.

Page 57 of the Fluke 1750 user manual — Regulatory Information for Wireless Communication
Page 57 · click to enlarge
Page 57

Power Recorder Regulatory Information for Wireless Communication

Regulatory Information for Wireless Communication

  • Any changes or modifications to the wireless 2.4 GHz radio not expressly approved by Fluke Corporation could void the user's authority to operate the equipment.
  • The distance between the Recorder and the human body must be ≥ 20 cm (7.9 inches) during wireless communication.
  • FCC ID: T68FLK1750 or QOQWT 11 (see rear decal on Fluke 1750 for the appropriate ID).

This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device.

Class A digital device: A digital device that is marketed for use in a commercial, industrial or business environment, exclusive of a device which is marketed for use by the general public or is intended to be used in the home.

Class B digital device: A digital device that is marketed for use in a residential environment notwithstanding use in commercial, business and industrial environments. Examples of such devices include, but are not limited to, personal computers, calculators, and similar electronic devices that are marketed for use by the general public.

The Fluke 1750 Power Recorder (wireless host) has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.

In addition, the internal limited modular radio has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:

  • Reorient or relocate the receiving antenna.
  • Increase the separation between the equipment and receiver.
  • Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
  • Consult the dealer or an experienced radio/TV technician for help.

The term “IC:” before the radio certification number only signifies that Industry of Canada technical specifications were met.

W Caution

Page 58 of the Fluke 1750 user manual — Specifications for the System: Recorder and Power Analyze Software
Page 58 · click to enlarge
Page 58

Specifications for the System: Recorder and Power Analyze Software

Power Quality Measurement Standards

Conformance IEC 61000-2-4:2002 Class 1, IEC 61000-4-30, IEEE519, IEEE1159,

IEEE1459, IEC 61557-12:2008, and EN50160

Clock/Calendar Leap years, 24-hour clock

Real-time Clock Accuracy Not more than ± 1 s/day

Internal Memory Capacity for Data At least 2 GB

Maximum Recording Period At least 31 days

Measurement Time Control Automatic

Maximum Number of Events Limited only by the size of the internal memory.

Power Requirements 100 to 240 V rms ± 10 %, 47-63 Hz, 40 W

Operating Time During Interruptions

(internal UPS operation) 5 minutes per interruption, 60 minutes total operating time without

Dimensions 215 x 310 x 35 mm (8.5 x 12.2 x 3.5 inch)

Mass (Weight) 6.3 kg (14 lbs)

Measurement Types One Phase Plus Neutral, One Phase IT No Neutral, One Phase Split

Phase, Three Phase Wye, Three Phase Delta, Three Phase IT, Three Phase High Leg, Three Phase Open Leg, 2 Element Delta, 2 ½ Element Wye

Input Channels Voltage: 5 channels, AC/DC Current: 5 channels

Voltage Channels Input resistance: 2 MΩ

Input capacitance<20 pF

Current Channels Self-identifying probes

Types availablecurrent clamps, Flexi-CTs

Measuring Method Simultaneous digital sampling of voltage and current. Digital PLL

synchronized sampling, internal frequency reference used during voltage drops.

PLL-Synchronization Source The PLL synchronizes to the A-N/L1-N voltage for wye power types,

and to the A-B/L1-L2 voltage for delta power types. All listed power types can be characterized as either wye or delta.

PLL Lock Range 42.5 to 69 Hz

Sampling Frequency Voltage and current: 256 samples/cycle

Inter-harmonics per IEC61000-4-72560 points / 10 cycles (50 Hz),

3072 points/12 cycles (60 Hz)

Transient Voltage5 MHz

A/D Resolution Voltage and current: 24 bits

Transient voltage14 bits

Voltage and Current Measurements

Voltage Measurement Range AC voltage: 1000 V rms ± 10 % Overrange

DC voltage± 1000 V + 10 % Overrange

Voltage Crest Factor 3 or less

Voltage Input Impedance 2 MΩ

Current Measurement Range Depends on current probe used

Current Crest Factor 4 or less

Current Input Characteristics 2 V rms = full scale, 1 MΩ Input Impedance for ferro CTs, low

Show 5 more labels / short text on this page
General Specifications · recharging · Input Specifications · Synchronization and Sampling · impedance for Flexi-CTs
Page 59 of the Fluke 1750 user manual — Voltage and Current Measurement Accuracy
Page 59 · click to enlarge
Page 59

Power Recorder Specifications for the System: Recorder and Power Analyze Software

Voltage and Current Measurement Accuracy

Measurement Type True rms calculated continuously: every cycle, every 1/2 cycle, and

every 10 or 12 cycles at 50 or 60 Hz respectively, as required by IEC 61000-4-30.

Measurement Uncertainty AC: ± 0.2 % reading ± 0.1 % full scale above 50 V rms

DC± 0.5 % reading ± 0.2 % full scale above 50 V dc

If the ac component of a dc signal is below 40 V rms, the dc uncertainty specification may have an offset.

Measurement Type True rms calculated continuously: every cycle, every 1/2 cycle, and

every 10 or 12 cycles at 50 or 60 Hz respectively, as required by standards

Measurement Uncertainty ± (0.1 % full scale + 0.5 % reading + current sensor accuracy, valid for

5 % to 100 % of current sensor range

Current Accuracy Ferromagnetic Clamps

± (0.1 % full scale + 0.2 % reading + current sensor accuracy), valid for 5 % to 100 % of current sensor range Flexible Current Probes ± (0.1 % full scale + 0.5 % reading + current sensor accuracy), valid for 5 % to 100 % of current sensor range

Measurement Type Waveshape sampling, not peak detect

Full Scale 8000 V pk

Sample Resolution 200 nS

Measurement Uncertainty ± 5 % reading ± 20 V (test parameters: 1000 V dc, 1000 V rms,

Dip (Sag) and Swell Measurements

Voltage Swell (rms swell)

Measurement Type True rms (one cycle calculation by overlapping each half cycle)

(voltage between lines is measured for 3P3W lines and phase voltage is measured for 3P4W lines)

Displayed Data Amplitude and duration of swell

Measurement Accuracy Same as rms voltage

Voltage Dip (RMS sag)

Measurement Type True rms (one cycle calculation by overlapping each half cycle)

(voltage between lines is measured for 3P3W lines and phase voltage is measured for 3P4W lines)

Displayed Data Amplitude and duration of dip or interruption

Measurement Accuracy Same as rms voltage

Measurement Type Same as Voltage Dip

Power and PF Measurement

Calculated per IEEE1459 for best performance when distortions exist

Measurement Type True rms calculated continuously: every cycle, and every 10 or 12

cycles at 50 or 60 Hz respectively, as required by standards

Measurement Accuracy ± 0.2 % reading ± 0.1 % full scale + current sensor accuracy

Measurement Range 42.5 to 69 Hz

Measurement Source Same as PLL synchronization source

Measurement Accuracy ± 10 mHz (10 to 110 % of range, with sine wave)

Accuracy ± 0.2% reading ± 0.1 % full scale + current sensor accuracy

Show 7 more labels / short text on this page
RMS Voltage · RMS Current · Transient Voltage (Impulse) · 100 kHz) · Voltage Dropout (Interruption) · Frequency · Reactive Power
Page 60 of the Fluke 1750 user manual — ETSI EN300328 (V1.7.1:2006)
Page 60 · click to enlarge
Page 60

Measurement Range -1.000 (leading) to 0.000 to +1.000 (lagging)

Measurement Accuracy ± 1 digit from the calculation of each measured value (± 3 digits for

Measurement Method Calculated from the phase difference between voltage fundamental

Measurement Range -1.000 (leading) to 0.000 to + 1.000 (lagging)

Measurement Accuracy ± 0.5 % reading ± 2 % full scale ± 1 digit

Voltage Unbalance and Phase Sequence

Measurement Method Positive sequence voltage divided by negative sequence voltage, per

IEC 61000-4-30

Harmonic Voltage and Current

Analysis Window rectangular

Analysis Order 1st to 50th order

Measurement Accuracy Voltage / Current: 1st to 20th orders: ± 0.5 % reading ± 0.2 % full scale

21st to 50th orders: ± 1 % reading ± 0.3 % full scale (current sensor accuracy must be included for current and power)

Measurement Method IEC 61000-4-7:2002

Inter-harmonic Voltage and Current (Intermediate Harmonics)

Analysis Window rectangular

Analysis Orders 0.5 to 49.5th order

Measurement Method IEC 61000-4-7:2002

Measurement Method as per IEC 61000-4-15:2003: 10 min (Pst), 2 h (Plt)

Connector RJ-45

Speed and Type 10/100 Base-T, auto MDIX

Communications Protocol TCP/IP over Ethernet

Connection wireless (2.4 GHz radio)

Speed up to 700 kbit/second

Communications Protocol Bluetooth SPP

Environmental and Safety Specifications

Operating Environment indoors or in covered area outdoors, up to 2000 m altitude

Storage Temperature and Humidity -20 °C to 50 °C, 80 % rh max, non-condensing

Operating Temperature and Humidity 0 °C to 40 °C, 80 % rh max, non-condensing

Maximum Rated Working Voltage

Voltage Terminals 1100 V rms

Voltage Durability 5550 V rms ac for 1 minute, between voltage input terminals, voltage

input terminals and current probes, and voltage input terminals and case (50/60 Hz, 1 mA sense current)

Enclosure Protection IP30 (per EN 60529)

EMC IEC 61326-1:2006 Class A

IEC 61000-3-2:2005, 3

IEC 61000-3-3:2005, 1.2 Edition

AS/NZS 4268:2008 (BT module only)

ETSI EN300328 (V1.7.1:2006)

Safety IEC 61010-1:2001, 2

Voltage input unitContamination Level 2

Overvoltage Category 1000 V CAT III, 600 V CAT IV (anticipated overvoltage: 8000 V)

Show 11 more labels / short text on this page
Power Factor · total) · Displacement Power Factor · and current fundamental · Voltage flicker · External Interface Specifications · LAN Interface · Wireless Controller Interface · Standards Conformance · rd Edition · nd Edition
Sources

References & links

Pages referenced in this manual — manufacturer, support and downloads.

Ask about the Fluke 1750

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Fluke 1750

Be the first to ask — other owners or our team will answer.