Thank You for purchasing the Fishman LDS9000
AirFreeTM dispenser. Our 9000-line of AirFree dispensers uses a patented
positive displacement technology to provide the highest level of repeatability
of any dispenser utilizing a syringe as a reservoir. We guarantee it.
Please spend a few moments familiarizing yourself with the LDS9000
dispenser. Our recommended testing procedures provide an interactive
tutorial on the dispenser’s functions.When complete, review the helpful
information we have included based on our 40 years of dispensing experience.
After completing the guide, should you have any additional questions please
don’t hesitate to contact Fishman or your authorized Fishman dealer.

T h e F i s h m a n P r o m i s e
We promise that our products will operate to your complete satisfaction. We strive for the highest standards regarding quality, workmanship and attention to detail.
If for any reason you feel that you are not satisfied with our products or support, I extend the offer to write or call me personally. Our interest is to build partnerships through understanding your needs and incorporating them into our products and services.Your trust in Fishman is our ultimate reward.


This manual contains all the information you will need to set up and test the 9000 series AirPulseTM dispensers.
The remainder of the manual applies to all 9000 series AirFreeTM
dispensers.
Unpack the shipping box and use the supplied checklist to assure all
items are present. If an item is missing, please contact us immediately.
Make sure the set up area has access to power. No compressed air will
be needed. You are about to experience the world’s only AirFreeTM fluid dispenser. No compressors, no filtration, no plumbing.
Activate your extended two (2) year no-fault warranty by following the
instructions in the enclosed “On Line Warranty” letter to activate online. Once your dispenser is set up, return to this section to familiarize yourself with the multiple functions of the program controls.
First Steps



The LDS9000 Dispensing System uses a patented Linear Drive System (LDS) to transfer force from the drive motor to the syringe piston. The motor, lead screw and syringe piston are linked mechanically. Each attachment (gun to reservoir, gun to piston, piston to reservoir) is specifically designed to provide zero flexing in the system during a dispense. Simply put, zero flexing is what makes the LDS9000 the world’s most repeatable dispenser.

Once the dispense cycle is complete, a programmed mechanical pullback retracts the drive rod which is mechanically attached to the AirFreeTM piston. The combined action of the plunger with the preprogrammed forward steps, and pullback feature produces a highly repeatable deposit with no post-dispense ooze. A pause between forward and reverse motion can be programmed in for use with compressible fluids.
The inner diameter and length of the syringe barrel are used to calculate dispense volume. For example, the inner diameter of a 10cc syringe barrel measures 1.63cm. One forward step of the motor moves the drive rod/piston .00127cm, resulting in .0026cc of fluid being displaced. With a 5cc barrel, it takes two steps to displace the same amount of fluid.
An electronic signal is sent via the electronic control cable that connects the dispense gun to the control box. The signal energizes the motor pushing the drive rod, attached to the piston, into the syringe barrel. The forward motion forces the fluid out to make a deposit.
The LDS9000 features a universal power supply. It automatically adjusts for input voltages ranging from 100 VAC to 240 VAC. There is no adjustment required by the user.
Flexible Control Cable · Power Supply


In manual mode, flip the Extend switch, located on the control panel, from Extend to Retract to engage the motor in reverse direction. This will draw fluid back into the reservoir stopping any ooze.
In both manual and auto mode, flip the Foot/Hand switch, from Foot to Hand to operate the dispenser with a finger switch. Finger switches are sold separately.
Manual/Auto Dispense · Reverse Motion—Drawback · Hand Control


1 VOLUME: 4.99
Wait 3-5 seconds for on-screen instructions
When in program mode, if you are not sure what to do wait approximately 3-5 seconds and rolling instructions appear. If you missed them, not to worry, they will repeat until you make an entry.
New or Saved Program
In Auto Mode hit the NEXT key. Hit the left arrow to enter a new program or hit the right arrow to enter a saved program.
Edit the existing program
In Auto mode hit the BACK key once to enter the program manager. Each time you hit the BACK key it shows a different program parameter. Find the one you wish to edit, make your edit and hit the ENTER key. Hit the NEXT key as many times as needed to return to program mode. Follow on-screen instructions.
In Auto mode when entering a new program or editing an existing one, when prompted for a dispense volume press the digit keys to type the amount you require. Press the ENTER key to save your input.
Skip forward/back through the menu
In Auto mode, press the NEXT key to skip one slot forward in the programming menu. Use the BACK key to skip back one slot in the programming menu.
On Screen Instructions · Enter a Volume

Erase an Input or Make a Selection
In Auto mode when entering a new program or editing, use the left and right arrow keys to erase and input or skip between choices such as multiple saved programs. Use the ENTER key to select a choice.
In auto or manual, always use the ENTER key to save an input or selection. The NEXT or BACK keys do not save—they only skip to the next choice. The ENTER key saves your data.
Jog Between Menu Choices
In auto or manual, always use the left or right arrow to move between menu selections.



SureLok™ Retaining Ring Cover


Holding the dispensing gun with the piston hub pointed upward, place the barrel over the piston hub of the dispense gun.
Holding both loosely. Step on the foot pedal until the barrel ears are raised at least 1/4” (.635mm) above the plate.

Attaching the Syringe (continued)

Manually twist the barrel in a clockwise direction until the luer lock connector is finger tight and the ears of the barrel are flush against the retaining plate. It may be necessary to set the EXTEND/RETRACT switch to RETRACT and step on the foot pedal to bring the barrel ears completely flush to the retaining plate. If you set the switch to RETRACT be sure to set it back to EXTEND before proceeding.
Slide the retaining ring cover over the syringe and rotate the cover in a clockwise direction.
The syringe should be firmly connected to the dispensing gun.


control unit will display a message: “Remove syringe and press enter.”
Twist the barrel reservoir clockwise to remove it from the piston hub.

The LDS9000 AirFreeTM Dispensing system allows 10 user-defined programs to be stored in memory, ranging from 0 to 9. The program stores:
Follow These Steps to Input a Program
3 5 10 30
Use the left and right arrow keys to select your syringe size.
.01 cc .00261-12.2335
Enter your dispense volume by using the keypad to enter your volume followed by pressing the ENTER key or use the left or right arrows to use the SpeedBarTM feature*.
.392 cc/sec .065 - 1.567
Enter your dispense rate by using the keypad to enter the volume followed by pressing the ENTER key or use the left or right arrows to use the SpeedBarTM feature*.
Enter the amount of back off by using the keypad followed by pressing the ENTER key or use the left or right arrows to use the SpeedBarTM feature*.
Enter the amount of delay (0-9 seconds) between forward and reverse motion by using the keypad followed by the ENTER key.
0 seconds .00 - 10.00
*SpeedbarTM—This feature is available in the volume input, dispense rate input and backoff input. Hold down the left or right arrow key to make this feature appear. Continue holding to make the bar increase or decrease. The longer you hold, the faster the bar moves. When satisfied, release the key and SpeedbarTM will show you your selected volume.
0cc 12cc

To save program, answer YES followed
by the program number (0-9) of the location you wish to save your program.
If you want the program to automatically load as the default program when the machine is turned on, answer YES to the default question.
<- YES NO ->
Follow These Steps to Edit a Program
the start of the program manager. The NEXT key moves you through each parameter to the bottom of the program manager.

Follow these steps to save a program to a different location.

program you will be asked if you wish or overwrite the program.
Follow these steps to restore a program at Start Up.
worry if you miss a program, the scroll feature will repeat until you make a selection
Follow these steps to restore a program while in an active program.


To dispense a continuous bead of fluid instead of a single dot, follow these steps.
Switch the AUTO/MANUAL switch to
MANUAL.

Set the shot size to the lowest setting.
For best results, set the dispense rate to slow, to minimize backpressure in the syringe.
As a beginning point, set drawback to 5 steps
Keep delay set to zero, then adjust as needed to eliminate any post dispense ooze.


First Shot Set Up
Suggested Tips




Avoid Air in the Syringe Barrel
Trapped air in the fluid creates voids that will make the fluid compressible. Although the LDS9000 Linear Drive System is stable enough to handle most compressible fluids excess air will force the user to use a greater amount of pullback. This will create a greater amount of forward and reverse motion in the syringe during a dispense cycle. Excess motion will put more stress on the fluid being dispensed. The goal is to reduce dispense motion as much as possible to protect the fluid and maximize repeatability. Paying attention to minimization during syringe filling will drastically improve performance.
For best results, use a Fishman barrel filing station (see next page). If needed, use a centrifuge or vacuum chamber to eliminate any trapped air.

Twist on a tip cap and pour your fluid into the barrel using a funnel. Insert a Venting Ribbon along the inside edge of the barrel, followed by an AirFree™ piston. Carefully press down until it contacts the fluid. While holding the piston in place gently pull the Venting Ribbon from the barrel.
Thick or Non Leveling Fluid
If the fluid comes packed in a caulking cartridge try loading the reservoir with a caulking gun. If not, spoon-feeding with a spatula often works. Follow the loading by inserting the Venting Ribbon along the inside of the barrel, followed by AirFree™ piston. Carefully press down to move the fluid to the bottom of the barrel and to remove trapped air. While holding the piston in place gently pull the Venting Ribbon from the barrel.
Pourable Fluid

For additional assistance, contact one of our Application Specialists:


Program mode is ideally suited for dot type applications or automated (XYZ system) beads. This mode allows you to program a desired volume of fluid for repeated and consistent output. In the program mode there are two utilities. Program Manager and the Save/Restore Manager.
The Program Manager has five adjustable parameters:
The selection of the proper barrel is critical to positive displacement dispensing. For the best results always select the smallest barrel reservoir that will work for your application. To minimize set up time we preprogrammed the control unit using standard Fishman luer lock barrels. Simply select the proper size (3,5,10,30) and move to the next parameter.
Having the ability to enter a dispense volume is what separates the LDS9000 from traditional pneumatic dispensing systems. Our patented technology allows us to apply a formula based on the syringe’s inner diameter and the distance traveled by the PosiLokTM
piston to arrive at the desired volume. This method drastically reduces the amount of time consumed by the guesswork of the pneumatic systems.
This is an important parameter that allows the LDS9000 to dispense a wide range of material. By altering the speed at which the Linear Drive System travels down the barrel, we can dispense non-compressible to thixotropic materials with little to no impact on rheology. The microprocessor calculates dispense rate in the cc per second.
Program Mode

Drawback:
Upon completion of the dispense cycle, fluid oozing out of the dispense tip is a common problem when dispensing compressible fluids. The Drawback parameter eliminates any oozing by implementing a mechanically reversed action at the end of the dispense cycle. Because the system is intelligent, it knows the distance traveled forward and the distance traveled back. Therefore, on the next dispense cycle, the system will add the distance traveled back to the forward motion, guaranteeing a positive displaced dispense every time.

Example: The system converts dispense volume to motor steps. Assume .02cc equals 20 steps. The following is what happens when 10 steps of reverse motion is added.
This parameter is most useful when potting or beading with a compressible fluid. A delay (in seconds) is applied between the end of the forward motion and the start of the reverse motion when drawback steps are programmed. This allows the entire programmed shot to dispense prior to the reverse motion.
The Save/Restore Manager has three adjustable parameters:
The Save/Restore Manager allows you to save and recall up to ten different programs, eliminating setup time when changing dispense volumes. Also, you may specify one program as the default program meaning this program will automatically load when the controller is turned on. This is useful when multiple LDS9000 are used on a production line. You are assured all systems will start with the same program.

Program is ideally suited for priming the dispense tip and manual dispensing such as potting and beads. In manual mode the parameters and function of the parameters are exactly the same as program mode. The only difference is that the system will only dispense fluid for the amount of time the user holds the foot pedal down.
The Linear Drive System
This unique design has made positive displacement with a barrel reservoir possible. Re-engineering the pneumatic piston concept, the high interference fit of the piston and the luer lock connection with the Linear Drive System enables intelligent dispensing. By increasing the interference between the O.D. of the piston and the I.D. of the barrel, we are able to completely seal the chamber eliminating material leakage. The luer lock connection provides complete control over forward and reverse motion. The intelligent control unit microprocessor knows the barrel I.D., the distance traveled forward and the distance traveled backward. With every dispense cycle the piston moves the exact distance every time, guaranteed. Drip-free dispensing, air pocket venting and all-plastic component dispensing are maintained with the new design.
The motor used inside the dispense gun is a linear actuator. The linear actuator creates forward motion by rotating nut between electrically charged poles. The drive screw runs through the nut located at the center of the motor. The design of the plastic gun housing captures the top of the drive screw preventing it from rotation. When electrically charged, the center nut will step from one magnetic pole to the other, moving the drive screw forward. It is important to note that the drive screw not rotating allows the attachment of the PosiLokTM piston, creating the linear motion critical to positive displacement. By reversing the charge, we are able to create reverse motion.
Manual Mode

Retaining Ring— The brawn of the AirFree dispensing technology
The multi-legged stainless steel design locks the syringe in place, making the syringe and gun one. In dispensing, any flexing between the barrel reservoir and surface to which it is secured represents a pressure loss and will change the amount of fluid dispensed. Due to the tight tolerance between the stainless steel retaining ring and the gun, there is no room for barrel flexing when the thrust-of-dispense cycle is applied. This is a key contributor to maintaining positive displacement dispensing.
The control unit gives the user control over volume output; dispense rate, reverse motion and motion delay. By imputing and manipulating these parameters, the microprocessor will calculate the number of steps needed to achieve the requested result. A universal switching power supply automatically adapts to multiple AC inputs. The foot pedal may be removed for automated activation using a dry contact.
Helpful Hints Priming the System
Removal of the tip cap and introduction of a new needle traps air pockets inside the stem of the syringe and hub of the needle. Because our system positively displaces fluid, it is very important no air is present in any portion of the syringe. Therefore, it is important that any trapped air is pushed out of the syringe prior to dispensing.
Same Size Barrels: Simply switch one barrel for the next. The only consideration is to prime the new dispense tip prior to dispensing (page ?). The LDS9000 recognizes barrels of the same size so it is not necessary to reset the system.
Different Size Barrels: When changing to a different size barrel, it is necessary to reset the lead screw, reset the dispense parameters and prime the dispense tip. To reset the lead screw remove the barrel reservoir, flip the Auto/Manual switch to Manual and the Extend/Retract switch to Retract. Hold the foot pedal down until the lead screw retracts all the way into the dispense gun.
Changing Barrels Reservoirs

With the power ON, here’s what happens when you press the foot pedal:
The program is engaged, the motor is energized beginning the dispense
cycle countdown.
The motor pushes the lead screw, with piston attached to the end,
down the barrel pushing the liquid out the dispensing tip.
At the end of the program, the motor is de-energized—stopping
forward movement of the lead screw and stopping all fluid flow.
If drawback is programmed, the motor is re-energized and begins a drawback cycle countdown.
The motor retracts the lead screw, with piston attached to the end,
creating a vacuum that stops any post dispense oozing. The vacuum is controlled by the software program.
When in manual mode the microprocessor switches to direct motor control by the foot pedal. As long as the foot pedal is pressed, the motor remains energized, providing a flow of liquid for filling or beading.

Problem: The fluid continues to ooze long after the dispense, (including drawback) is completed.
Potential Solutions:
Problem: The fluid is blowing by the AirFreeTM piston.
Potential Solutions:

FOOT Setting for using a foot pedal to activate dispensing cycle
HAND Setting for using finger switch to activate a dispense cycle
SYR Syringe— barrel reservoir selection
DIA Length—user-defined syringe inside length from ears to bottom in
V Volume—Volume to be dispensed, in cubic centimeters
R Dispense Rate—The rate, in cubic centimeters per second, at
which you want to dispense the fluid
PRGM Program key to enter or exit at AUTO DISPENSE PROGRAM
ENTER Key to activate desired parameter
NEXT Key to advance to the next screen without changing the exiting
SYRINGE ATTACHMENT The ideal position for the ears of the barrel reservoir after locking
it into place, is flush up against the retaining plate. Note: Always attach a new barrel reservoir with the tip cap on.
PRIMING THE SYSTEM Eliminates any trapped air between the barrel reservoir and
needle. Always prime the system at start up.
PARAMETER LIMITS If you want to know the min., max, or high and low values of
parameters for a given syringe, type in 999 at the prompt and press the ENTER key. View the display; it will flash the values on the screen and then return.
RETRACT When you are in program mode and you change the
EXTEND/RETRACT switch to RETRACT, the system will prompt you to hit the ENTER key or flip the switch back to EXTEND. This is a safety precaution. If you retract with the PosiLok
TM piston attached, the barrel reservoir will fill up with air from the open end. Before you use the retract feature, be sure you remove your barrel reservoir. When retracting the lead screw in manual mode, all you have to do is flip the EXTEND/RETRACT switch to RETRACT. The PosiLokTM
piston will retract for as long as the foot pedal is depressed.
millimeters · parameters

DISPENSE RATE Always start with the slowest recommended rate and increase
incrementally.
DRAWBACK DELAY When using a small needle you can take advantage of the
drawback delay feature to give you more control over your output volume. This feature allows the fluid to flow out of the syringe before you drawback the PosiLokTM piston and begin your next dispense cycle.
FACTORY SETTINGS The factory settings will be replaced with your settings when you
save into location #1 and #2
SAVE When you save an application into a location that has a
previously stored application, you will be asked to replace the existing application.
RESTORE When you restore an application from a location, the parameters
will flash on the display. You will then be asked to restore the application.
ACTIVATING A PARAMETER if you notice that a parameter that you set didn’t take effect,
make sure you pressed the ENTER key before proceeding to the next screen. If you did not press the ENTER key, the values you entered will not take effect on the next dispensing cycle.
NEEDLE SELECTION When selecting a needle we recommend using the FreeFlowTM tip
with the LDS9000. This tip allows the least amount of flow restriction thus allowing positive flow of the fluid from the syringe.
PREPACKAGED SYRINGE Must be free of air pockets to receive the full benefits of the
LDS9000. Contact your Fishman representative for AirFreeTM
filling suggestions.

The charts below are only a small sample of available Fishman components. Please refer to the complete Fishman catalog to view the entire product line.
PosiLok TM Flexible Dispense Tips
The unique multi-legged design firmly locks the barrel into place eliminating any possibility of pressure loss during dispense.
Soft polypropylene burr and flash free tips assure accurate deposit control.
Gauge ID .5” 1.5” Color 15G .049” Z532115 Z534115 Brown 18G .032” Z532118 Z534118 Dk. Green 20G .019” Z532120 Z534120 Pink 25G .014” Z532125 Z534125 Red
Size Retaining Ring Quantity Per Pack 3cc O312291 1 10cc O314291 1

Accessories
Each package contains same amount of barrels and PosiLok™engageable pistons.
Gun Stand Holds the dispense gun when not in use.
Size Clear Quantity Per Pack 3cc O502051 50 10cc O504171 30
Piston Stabilizer Attaches to the end of the Linear Drive System dispense rod. Clam shell design clamps onto the piston hub. Each kit contains a piston stabilizer, assembly tool and instructions
5cc Stabilizer Kit Y010101 10cc Stabilizer Kit Y020101 30cc Stabilizer Kit Y030101
ID 1/2” Length Hub Color
.020” Z592121 Gray .006” Z592125 Pink
Automation Gun Bracket Mounts the gun to an XYZ dispense station.
LDS9000 Stand D600001 · LDS9000 Bracket Y200001



Fishman Corporation (Fishman) warrants its product to be free from defects in materials and workmanship for a period of two years from the date of purchase.
Under this warranty, Fishman will repair or replace any defective part or, at its own option, the complete dispenser regardless of fault.To obtain service under warranty, product must be returned, freight prepaid, to Fishman.
Before use, the customer must determine the suitability of the product for their intended use and therefore customer assumes all risk and liability connected with products use.
At no time under any circumstance shall any liability or obligation of Fishman arising from the warranty exceed the purchase price of the product.
Except as provided herein, Fishman makes no warranties, express or implied, including warranties of merchantability and fitness for a particular purpose. In no event will Fishman be liable for direct, indirect, special, incidental, or consequential damages arising out of use of this product.
O
R
N
D
U
S
T
E
D
R
A
O
N
M
A
D
E
N
S
U
S
A
M
S
A
M
N
M
U
F
O
A
T
A
S
R
U
U
N
N
E
D
A
M