©2009, Micro Motion, Inc. All rights reserved. Viscomaster is a registered trademark, and Viscomaster Dynamic is a trademark of one of the companies of Emerson Electric Co. Micro Motion is a registered trade name of Micro Motion, Inc., Boulder, Colorado. The Micro Motion and Emerson logos are trademarks and service marks of Emerson Electric Co. All other trademarks are property of their respective owners.
Micro Motion pursues a policy of continuous development and product improvement. The specification in this document may therefore be changed without notice. To the best of our knowledge, the information contained in this document is accurate and Micro Motion cannot be held responsible for any errors, omissions, or other misinformation contained herein. No part of this document may be photocopied or reproduced without prior written consent of Micro Motion.

Chapter 1 Introduction 1 1.1 Safety guidelines 1 1.2 About the meter 1 1.2.1 What is it? 1 1.2.2 7829 Viscomaster meter measurements 2 1.2.3 7829 Viscomaster Dynamic meter measurements 2 1.2.4 What is it used for? 3
Chapter 2 Installation 5 2.1 Introduction 5 2.2 Installation effects 6 2.2.1 Boundary effects 6 2.2.2 Fluid at the sensor 8 2.2.3 Thermal effects 8 2.2.4 Entrained gas 9 2.2.5 Solids contamination 9 2.2.6 Vibration effects 9 2.3 General fitting notes 9 2.4 Standard installations 10 2.4.1 Overview 10 2.4.2 Meter orientation 11 2.4.3 Flow-through chamber installation 12 2.4.4 VAF Viscosense retrofit 13 2.4.5 VAF Viscotherm retrofit 16 2.5 Commissioning 18 2.6 During normal running 18 2.7 Removal and refitting procedure 18
Chapter 3 Electrical Connections 21 3.1 Introduction 21 3.2 Installation considerations 22 3.2.1 Power supply 22 3.2.2 EMC 22 3.2.3 Ground connections 23 3.2.4 Cabling requirements 23 3.2.5 Surge protection 23 3.2.6 Installation in explosive areas 24 3.3 Wiring the meter 26 3.4 Power supply input 26 3.5 Modbus (RS-485) 27 3.6 4-20 mA outputs 28 3.7 Wiring procedure 30 3.8 Further information on RS-485 32 3.8.1 RS-485 32 3.8.2 RS-485 to RS-232 32
Installation and Configuration Manual i

3.8.3 RS-485 multi-drop 35 3.8.4 Transmission mode 35
Chapter 4 Using ADView and ProLink II 37 4.1 Using ADView software 37 4.1.1 What is ADView? 37 4.1.2 Installing ADView 37 4.1.3 Starting ADView 38 4.1.4 Understanding ADView features 40 4.2 Using ProLink II software 45 4.2.1 Overview 45 4.2.2 Requirements 45 4.2.3 Connecting from a PC to a meter 45 4.2.4 ProLink II configuration upload/download 46 4.2.5 ProLink II language 46
Chapter 5 Calibration Check 47 5.1 Calibration 47 5.1.1 Factory calibration 47 5.1.2 Calibration of Transfer Standards 47 5.1.3 Instrument calibration 48 5.1.4 General viscosity equation 48 5.1.5 General density equation 48 5.2 Calibration certificate examples 49 5.2.1 Viscomaster sample calibration certificate 49 5.2.2 Viscomaster Dynamic sample calibration certificate 50 5.3 User calibration checks 51 5.3.1 Ambient air calibration check 51 5.3.2 On-line calibration adjustment 51
Chapter 6 General Maintenance 53 6.1 Overview 53 6.2 General maintenance 53 6.2.1 Physical checks 53 6.2.2 Electrical check 54 6.2.3 Performance check 54 6.2.4 Calibration check 54 6.3 Fault analysis and remedial action 54 6.3.1 Troubleshooting faults 54 6.3.2 Mechanical servicing 56 6.3.3 Time period trap 56
Appendix A Factory Default Settings 59
A.1 Default configuration for analog outputs 59
Appendix B Calculated Parameters 61
B.1 Overview 61 B.2 Base density referral 61 B.2.1 API density referral 61
ii Micro Motion 7829 Viscomaster® and Viscomaster Dynamic™ Viscosity Meters

B.3 Kinematic viscosity 63 B.4 Base kinematic viscosity referral using ASTM D341 63 B.5 Ignition quality 63
Appendix C Safety Certification 65
C.1 Safety certification 65
Appendix D Modbus Communications 67
D.1 Overview 67 D.2 Accessing Modbus registers 68 D.2.1 Establishing Modbus communications 68 D.3 Modbus implementation 68 D.3.1 Register size and content 68 D.4 Modbus register assignments 69 D.5 Index codes 72 D.5.1 API product type 73 D.5.2 Pressure, Temperature, Density and other Units 73 D.5.3 Output averaging time 73 D.5.4 Analog output selection 74 D.5.5 Referral temperature 74 D.5.6 Software version 74 D.5.7 Hardware type 74 D.5.8 Unit type 75 D.5.9 Status register flags 75 D.5.10 Line dynamic viscosity units 76 D.6 Establishing Modbus communications 76 D.7 Example of direct Modbus access 78 D.7.1 Example 1: Reading line density (16-bit register size) 78 D.7.2 Example 2: Reading line density (32-bit register size) 79
Appendix E Product Data 81
E.1 Density / temperature relationship of hydrocarbon products 81 E.1.1 Crude oil 81 E.1.2 Refined products 81 E.1.3 Platinum resistance law 82 E.1.4 Density of ambient air 83 E.1.5 Density of water 83 E.1.6 Velocity of sound in liquids 84
Appendix F Return Policy 87
F.1 General guidelines 87 F.2 New and unused equipment 87 F.3 Used equipment 87
Installation and Configuration Manual iii

iv Micro Motion 7829 Viscomaster® and Viscomaster Dynamic™ Viscosity Meters

Installation Using ADView and ProLink II Electrical Connections Introduction
Handle the 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter with great care.
Safety messages are provided throughout this manual to protect personnel and equipment. Read each safety message carefully before proceeding to the next step.
1.2.1 What is it?
The 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter is a digital viscosity meter, based on the proven tuning fork technology of Micro Motion. It is an all-welded sensor designed to be mounted directly into a pipeline or in a tank. Viscosity and density are determined from the resonance of the tuning fork immersed in the fluid, and a temperature sensor (RTD) is also fitted within the meter.
The meter is available in a 316 stainless steel, and the immersed tines can be laminated with PFA to inhibit the build up of residues such as asphaltenes.
The meter contains integral processing electronics to provide full configuration, enabling it to perform a variety of calculations.

Two forms of output are available:
No signal converter is required, which simplifies wiring and enables the meter to be connected directly to a plant monitoring and control systems and/or a local indicator.
The meter is factory set to perform API density referral. Re-configuration of the meter’s default settings (see Appendix A) is achieved by linking a PC to the Modbus (RS-485) connection and running Micro Motion's ADView or ProLink II (v2.9 or later) software. Once configured, the PC can be removed.
The 7829 Viscomaster meter directly measures the following fluid properties:
From these properties, the meter calculates:
The 7829 Viscomaster Dynamic meter directly measures the following fluid properties:
From these properties, the meter calculates:

1.2.4 What is it used for?
The 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter is designed specifically to control the viscosity of Heavy Fuel Oil (HFO) used by power plants. This is typically achieved by adjusting the heating of incoming HFO to maintain the viscosity within the limits set by the engine manufacturer.
Installation Using ADView and ProLink II Electrical Connections Introduction
HFO is a low cost, high viscosity fuel derived from refinery wastes. The quality of the oil and its viscosity/temperature characteristics can vary due to:
HFO heating is usually required to ensure that the viscosity of the oil at the injectors of a large diesel engine or burner nozzle is maintained at the optimum value (typically between 10 cSt and 20 cSt). Failure to observe the viscosity limits results in inefficient combustion, pollution problems and higher operating costs (either due to excessive fuel being burnt or premature wear to engine components).
Since simple temperature control has been shown to be ineffective due to the variability of oil quality in HFO, viscosity control is usually performed.
A typical HFO fuel circuit is shown in Figure 1-1.

Figure 1-1 Typical HFO Fuel Circuit

1 HFO fuel tank 2 Mixer 3 Re-circulation pump 4 Fuel heater 5 Control valve 6 Viscomaster
7 Heater Controller 8 Personal Computer (PC) 9 Engine Management Computer 10 Burner or 11 Large Diesel Engine 12 Unburned fuel return

Installation Using ADView and ProLink II Electrical Connections Introduction
All drawings and dimensions given in this manual are given here for planning purposes only. Before commencing fabrication, reference should always be made to the current issue of the appropriate drawings. Contact Micro Motion for details.
For further information on handling and using the meter, see “Safety guidelines” on page 1
There are a variety of external factors that affect the ability of the 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter to operate successfully. In order to ensure that your system works correctly, the effects of these factors must be taken into consideration when designing your installation.
There are two main aspects to consider:
Factors which may adversely affect accuracy and repeatability include:
In some applications, absolute accuracy is less important than repeatability. For example, in a system where the control parameters are initially adjusted for optimum performance, and thereafter only checked periodically.
The term achievable accuracy can be used to describe a measure of the product quality that can be realistically obtained from a process system. It is a function of measurement accuracy, stability and system response. High accuracy alone is no guarantee of good product quality if the response time of the system is measured in tens of minutes, or if the measurement bears little relevance to the operation of the system. Similarly, systems which require constant calibration and maintenance cannot achieve good achievable accuracy.

Factors which may adversely affect the relevance of the measurements could include:
Unlike other Micro Motion meters, the vibrating tines of the 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter are not totally enclosed. The walls of the pipe or tank in which the meter is installed will introduce boundaries to the fluid flow, and this will have an effect on the calibration of the sensor.
To overcome this, Micro Motion calibrates the meter under a variety of pre-defined conditions corresponding to the installation and pipe schedule. This condition is selected when ordering the 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter, so that by calibrating the meter under the same boundary conditions as the installation, the need for additional on-site calibration is eliminated.
Any insertion device or meter can only measure the properties of the fluid within the region of fluid to which it is sensitive.
For practical reasons, it is helpful to consider the sensitive, or effective region, for the viscometer as an ovoid centered on the tips of the tines with its long axis aligned with the direction in which the tines vibrate, as shown below. The meter is insensitive to the properties of the fluid outside this region and progressively more sensitive to fluid properties the closer the fluid is to the tines. Density can be considered a “mass centered” effect and viscosity a “surface centered” effect in this visualization; i.e. the measurement of density is more uniformly sensitive to the density of fluid throughout the region while viscosity measurement is much more critically sensitive to fluid on the surface of the tines.

If part of this volume is taken up by the pipework or fittings there is said to be a boundary effect; i.e., the intrusion of the pipe walls will alter the calibration. The diagram below illustrates the meter installed in a pocket on the side of a 4" (100 mm) horizontal pipe line (viewed from above). The effective region is completely enclosed within the pipe line and thus is completely fluid.


Installation Using ADView and ProLink II Electrical Connections Introduction
This next view shows other pipe outlines superimposed:

The smaller circle represents a 4" (100 mm) vertical pipe, which because the meter orientation is constant irrespective of pipe orientation intersects the effective region. The 6" (150 mm) pipe is the smallest pipe diameter to completely enclose the effective region when the pipe is vertical. Thus smaller pipe diameters can lead to a variety of different geometries which would each require a separate calibration.
An alternative condition is shown in the next diagram where the side pocket is extended until it passes completely through the effective region producing a “core”:


From this, it would appear that almost every installation requires a separate in situ calibration – a very undesirable situation. The problem is resolved by providing standard calibration geometries which can be used in all pipe work configurations and thereby allow the factory calibration conditions to be reproduced in the process.
The fluid in the effective zone of the meter must be of uniform composition and at uniform temperature. It must be representative of the fluid flow as a whole.
This is achieved either by mixing of the fluid either using a static inline mixer or taking advantage of any natural pipe condition that tends to cause mixing, such as pump discharge, partially open valves. The viscometer should be installed downstream where the flow is just returning to laminar flow conditions.
Avoid temperature gradients in the fluid and in the pipe work and fittings immediately upstream and downstream of the viscometer.
Always insulate the viscometer and surrounding pipework thoroughly. Insulation must be at least 1" (25 mm) of rockwool, preferably 2" (50 mm) (or equivalent insulating heat jacket) and enclosed in a sealed protective casing to prevent moisture ingress, air circulation, and crushing of the insulation. Special insulation jackets are available from Micro Motion for the flow-through chambers, which, because of the low volumetric flow rates and hence low heat flow, are more vulnerable to temperature effects.
Avoid direct heating or cooling of the viscometer and associated pipe work upstream and downstream that is likely to create temperature gradients. If it is necessary to provide protection against cooling due to loss of flow, electrical trace heating may be applied, provided it is thermostatically controlled and the thermostat is set to operate below the minimum operating temperature of the system.

2.2.4 Entrained gas
Gas pockets can disrupt the measurement. A brief disruption in the signal caused by transient gas pockets can be negated in the signal conditioning software, but more frequent disruptions or serious gas entrainment must be avoided. This can be achieved by observing the following conditions:
Installation Using ADView and ProLink II Electrical Connections Introduction
The 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter has been extensively tested under severe vibration conditions, both in the test laboratory and Marine/Power Station/Burner applications. The meter is approved according to the Lloyds Register standard, levels ENV 1, 2 and 3 and operates correctly up to the classification level of ENV4 (vibration test 2). This vibration level, ENV 4 includes correct operation at vibration levels of 4 g rms between frequencies of 5–100 Hz, and is used to describe the requirements for engine mounted equipment.
If vibration levels exceed these limits, or the meters are not installed as recommended by Micro Motion, Micro Motion cannot take responsibility for the correct operation of these units.
The 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter uses a 1.5” Swagelok style of fitting which requires no seals, minimizing maintenance and spares. These fittings are leak proof over a wide range of pressure and temperature conditions, and during rapid temperature cycling, which may occur during the transfer from HFO to distillate fuel.
The meter should normally be installed horizontally, with the slot between the tines vertical; this ensures that, for low flow rates, any solids or gas bubbles are not trapped. When installed in a flow-through chamber, however, provided that the flow rate is within the recommended range, the transmitter can be mounted horizontally or vertically.
Allow at least 7.8” (200 mm) clearance to enable the meter to be removed from the fitting.

2.4 Standard installations
2.4.1 Overview
To overcome the need for in situ calibration for every installation, three standard installations are proposed. If an installation conforms to one of these standards, the factory calibration of the 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter is valid, and in-situ calibration unnecessary. The three installations are summarized in Table 2-1.
Standard installation
VAF VISCOTHERM / NAKAKITA Retrofit
VAF VISCOSENSE Retrofit Flow-through chamber
Viscomaster tines project into adapter kit with 2½“ Schedule 40 boundary.
Viscomaster tines are contained in a side pocket off the main flow, recessed by 25.4 mm (1 inch).
Viscomaster tines are contained in a flow-through chamber in which fluid is circulated from the main flow.
10 to 330 l/min (0.6 to 20 m3/hr) (2.6 to 87 US gal./min).
10 to 330 l/min (0.6 to 20 m3/hr) (2.6 to 87 US gal./min).
10 to 330 l/min (0.6 to 20 m3/hr) (2.6 to 87 US gal./min).
Up to 100 cSt. Up to 100 cSt Up to 100 cSt.
-50 to 200°C (-58 to 392°F).
-50 to 200°C (-58 to 392°F).
-50 to 200°C (-58 to 392°F).
Main flow pipe size
As defined by capillary Viscomaster chamber.
50 mm (2"). 50 mm (2").
(1) Viscomaster tines project into adapter kit with 2-½“ Schedule 40 boundary and retracted by 1” (25 mm).

These three types of standard installation are graphically shown in the following schematic.
Installation Using ADView and ProLink II Electrical Connections Introduction

Key 1 In-line viscometer replacement 2 Capillary viscometer replacement 3 Micro Motion flow through chamber
The meter must always be installed horizontally, and orientated to allow flow in the gap between the tines. This is irrespective of the pipe line orientation, and helps to prevent the trapping of bubbles or solids on the meter.

Figure 2-1 Meter orientation

Flow-through chambers are fabricated by Micro Motion, and are available with either weld prepared ends or with flange or compression fittings for connection into the process pipe lines. They are available with 2" NB inlet and outlet pipes.
Conditions:
[-40 °C to 200 °C (-40 °F to 392 °F) in hazardous areas]
The PT100 is a direct insertion type, without a thermowell, and uses a ¾" Swagelok connection.
The diagram below shows an example of this type of standard installation.


Installation Using ADView and ProLink II Electrical Connections Introduction
The three compression fittings on the flow pockets (½" drain, ¾" temp probe, and 1-½" mounting nut for the meter) are rated to above the working pressure of the flow pocket. The fittings may be Swagelok or Parker; both are used in manufacture.
The fittings are certified to the following standards:
Conditions:
This retro-fit kit has been specifically designed to provide a simple, direct replacement for existing vibration-type viscometers. Typically, the flange-to-flange distance is 5.9” (150 mm), although other larger versions can be accommodated (contact Micro Motion for details of the Universal retro-fit adapter). Usually, no pipework changes are necessary.

The 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter is mounted 0.98” (25 mm) away from the main flow line, allowing good product mixing, sensor protection and stable measurement conditions.
Typical dimensions are shown in Figure 2-2.

Figure 2-2 VAF Viscosense retrofit dimensions

Installation Using ADView and ProLink II Electrical Connections Introduction

2.4.5 VAF Viscotherm retrofit
Conditions:
This retro-fit kit has been specifically designed to provide a simple, direct replacement for existing capillary viscometers. Typically, these viscometers are designed to operate with their own measurement chamber, to which this adapter will be attached. No pipework changes are necessary.
Typical dimensions are shown in Figure 2-3.

Figure 2-3 VAF Viscotherm retrofit dimensions

Installation Using ADView and ProLink II Electrical Connections Introduction

2.5 Commissioning
Observe and record the normal operating temperatures and viscosity readings. You can monitor the system using ADView or ProLink II. (See the Using ADView and ProLink II chapter.)
When several systems are run in parallel and use the same fuel source, comparison of the readings between installations can be a useful indicator of possible system faults. Differences between readings or changes from the normally observed conditions should always be investigated to confirm that instrumentation is functioning correctly.
Particular attention should be paid to the conditions before and after engine shutdowns in order to detect any possibility of asphaltenes coating (precipitation of asphaltenes from the HFO caused by dilution with distillate fuel) which may cause the instrument to read high. If the re-circulation flow is high enough or the instruments have been supplied with PFA coating, asphaltenes or any other deposits should quickly be removed and the expected operating temperatures should be restored.
If the meter is still reading high and the oil quality is known not to have changed, then the instrument should be removed and cleaned with a rag. Removal should only be performed in accordance with the engine or burner manufacturers’ recommendations or in accordance with safe site practice. This must include isolation and depressurization.
All national and international safety regulations should be observed.
Observe safe working practice, wear protective clothing and safety glasses, and use suitable gloves to prevent burns or absorption of hot oil.
Check that the isolation valves have been fully closed, remove insulation and allow to cool to a safe level (cooling will tend to reduce any retained pressure) and de-pressurize the system if a drain valve or pressure relieving valve is fitted.

When the above conditions are satisfied, slacken the lock nut by 1-½ to 2 turns, sufficient for the sensor to be rocked. (If necessary, jolt the meter loose with a blow of the hand to the amplifier housing.) This will allow the seal between the sensor and the chamber retro-fit kit to be broken. Do not slacken the lock nut further unless the seal is broken and the sensor is definitely loose in the fitting.
Installation Using ADView and ProLink II Electrical Connections Introduction
Note: If the system is still pressurized, the meter may lift and be held against the retaining nut. Rocking and alternately pushing the sensor in and out of the pocket within the limits allowed by the slackened nut will break any seal and allow oil under pressure to seep past the lock nut. If this leakage is excessive, re-tighten the lock nut and take further action to de-pressurize the system.
When the meter can be rocked in the flow chamber and there is no serious or continuous escape of oil, it is safe to remove the lock nut.
Always keep all parts of your body away from the axis of the sensor (i.e., the direction in which the sensor will be withdrawn). If the system is under pressure or suddenly comes under pressure (e.g., due to valve failure or pump start), and the lock nut is not in place, the instrument may be forcibly ejected from the flow chamber and cause serious injury.
Clean and maintain the meter as directed and then refit it, as described in Section 2.5.


Chapter 3 Electrical Connections
Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction
For installations in hazardous areas:
The 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter has two types of output:
The Viscomaster Dynamic has a single fully configurable output proportional to a user-specified parameter. The Viscomaster has two fully configurable outputs.
The parameters that can be output on each analog output are as follows:
Analog Output 1 Analog Output 2 Analog Output 1 Analog Output 2
Dynamic viscosity (cP) Temperature (1)
Dynamic viscosity (cP) Dynamic viscosity (cP)
Kinematic viscosity (cSt) (1) Kinematic viscosity (cSt) (1) Kinematic viscosity (cSt)
Line density Temperature Temperature (1)
Temperature Line density Line density
Referred density Referred density
Referred viscosity Referred viscosity
CCAI CCAI
(1) Factory default selection.

It is recommended that both outputs are installed, requiring a minimum of eight wires (two for each output, and two for power). Although you may not immediately require the Modbus connection, it may be required for in-situ calibration adjustment and future system enhancements, and the cost of the additional wires is trivial compared to the expense of installing them retrospectively.
A number of factors must be taken into account when planning the electrical installation. These include:
3.2.1 Power supply
The power supply to the 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter must have the following requirements:
If several meters are to be used within a local area, one power supply can be used to power them all; where the meters are distributed over a wide area and cabling costs are high, it may be more cost effective to use several smaller, local power supplies.
Upon leaving the factory, the two 4-20 mA analog outputs are non-isolated as they are powered through internal links to the power supply input. However, if split-pads “LNK A” (Analog Output 1) and “LNK B” (Analog Output 2) by the terminal block are ‘broken’, they become isolated and require a separate 20-28 VDC power supply (see the 4–20 mA outputs section for details).
If an RS-232 to RS-485 converter is used (for example to connect to a serial port on a PC), this may also require a power supply (see the Further information on RS-485 section for details).
Care should be taken where there is the possibility of significant common-mode voltages between different parts of the system. For example, if the meter is locally powered from a power supply which is at a different potential to the RS-485 ground connection (if used).
To meet the EC Directive for EMC (Electromagnetic Compatibility), it is recommended that the meter be connected using a suitable instrumentation cable containing an overall screen. This should be earthed at both ends of the cable. At the meter, the screen can be earthed to the meter body (and therefore to the pipework), using a conductive cable gland.

3.2.3 Ground connections
It is not necessary to earth the meter through a separate connection; this is usually achieved directly through the metalwork of the installation.
Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction
The electronics and communications connections (RS-485/Modbus and 4-20 mA analog output) of the meter are not connected to the body of the meter. This means that the negative terminal of the power supply can be at a different potential to the earthed bodywork.
In the majority of applications, it is not necessary to connect the RS-485 ground connection. In areas where there is a significant amount of electrical noise, higher communications integrity may be obtained by connecting the negative power terminal (pin 2) of the meter to the communications ground. If this is done, it is important to ensure that the possibility of ground loops (caused by differences in earth potential) is eliminated.
Although it is possible to connect separate cables to the meter for power, RS-485 and the 4-20 mA analog output, it is recommended that all connections are made through one instrumentation-grade cable.
Connections for the Analog and Modbus signals should be individually screened twisted-pairs with an overall screen, foil or braid for the cable. Where permissible, the screen should be connected to earth at both ends. (At the meter, this is best done using a conductive cable gland.)
Cables should conform to BS2538. In the USA, use Belden 9402 (two-pair) or Beldon 85220 (single-pair). Other cables that are suitable are those that meet BS5308 Multi-pair Instrumentation Types 1 and 2, Belden Types 9500, 9873, 9874, 9773, 9774 etc.
The typical maximum recommended cable length for the above cable types is 1000 m (3200 ft), but care must be taken to ensure that the power supply at the meter is at least 20 V. Thus, for 24 V power supply, the overall resistance for the power supply connections (both wires in series) must be less than 100 ohms.
In order to complete the wiring, you will need the following parts:
The gland adapter and blanking plug are supplied with the meter – these two parts are Exd rated. However, you will need to get a suitably rated cable gland:
In hazardous areas, all parts must be explosion-proof. Alternative parts may be required in order to meet local electrical installation regulations.
Careful consideration should be given to the likelihood of power supply surges or lightning strikes. The power supply connections of the meter have a surge arrestor fitted that gives protection against power supply transients.
If there is a possibility of lightning strikes, external surge protection devices - one for each pair of signals and the power supply - should be installed as close to the meter as possible.

Another method of surge protection is to connect an MOV (Metal Oxide Varistor) (breakdown voltage >30 V) with an NE-2 neon bulb in parallel across each wire and ground. These can be mounted in a junction box close to the meter.
If the RS-485/Modbus output is permanently connected to a PC, an independently powered, fully isolated RS-485 to RS-232 converter should be used. (See the Further information on RS-485 section for details).
For installations in hazardous areas:
The meter is an explosion-proof and flameproof device. However, it is essential to observe the rules of compliance with current standards concerning flameproof equipment:

Figure 3-1 Wiring diagram
Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction
Power supply (20...28 Vdc at 50 mA)
RS-485 A
To RS-232 port on a PC running ADView or ProLink II (v2.9 or later) software for monitoring, maintenance and configuration.
4-20 mA output 1 +
4-20 mA output 1 -
Passive outputs (see the 4-20 mA outputs section for more information)
4-20 mA output 2 +
8 Analog o/p +
4-20 mA output 2 - Analog o/p -

3.3 Wiring the meter
Figure 3-2 shows the terminal board of the 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter. To reveal the terminal board, it is necessary to unscrew the housing cap; the procedure is described in the Wiring Procedure section.
The connections to the meter are:
It is recommended that you install all connections (eight cores) at installation, to avoid the possibility of expensive alterations to the cabling at a later date. Typically, four pairs of shielded 19/0.30 mm2 (#16 AWG) to 19/0.15 mm2 (#22 AWG) wires are used.

Terminals 1 and 2 are for connecting an external 24 VDC power supply, as guided in Figure 3-3.
Ensure that the loop resistance of the cable(s) is such that the voltage at the meter terminals is greater than 20 volts. (The maximum voltage at the meter terminals is 28 VDC.)

Figure 3-3 Power supply connections

Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction
Terminals 3 and 4 are for RS-485/Modbus connections to a PC, as shown in Figure 3-4. For cable distances above 100 m, see the Further information on RS-485 section.
The RS-485/232 converter and PC are not normally installed permanently. However it is strongly recommended that the wiring to the meter is made at the time of installation.
For detailed information on RS-485, see the Further information on RS-485 section.

Figure 3-4 Modbus connections < 100 m


9-pin DIN connector version
Terminals 5, 6, 7 and 8 are for connecting the two 4-20 mA analog outputs to external devices, such as a signal converter. Upon leaving the factory, the two 4-20 mA analog outputs are non-isolated as they are powered through internal links to the Power Supply Input.
Terminal block version

Figure 3-5 4–20 mA output using the main power supply

Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction
However, if split-pads “LNK A” (Analog Output 1) and “LNK B” (Analog Output 2) by the terminal block are ‘broken’, they become isolated and require direct connections to another external 20–28 VDC power supply. A second or third external 20–28 VDC supply can be used.
To isolate the analog outputs from internal power, use a sharp knife to cut the fine metal strip (or trace) for the appropriate split-pad (see Figure 3-6).
Figure 3-6 Isolating an analog output from internal power (for external power connection)

Non-isolated analog output (default) Connected to internal power (split-pad with trace)
Isolated output Disconnected from internal power for external power connection (split-pad with broken, or cut, trace)
Location of LNK A and LNK B split-pads
Example split-pads

Figure 3-7 4–20 mA output using a third power supply

Fault conditions within the meter are indicated by a 2 mA output. If this is detected, the Modbus link can be used to interrogate the meter to establish the likely cause of the problem.
GRUB SCREW



Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction
a: ¾” NPT Blanking Plug. b: ¾” NPT to M20 adaptor. c: M20 cable gland.


TIGHTEN CAP

3.8 Further information on RS-485
3.8.1 RS-485
The meter’s Modbus communications uses the RS-485 electrical standard. This uses the difference between the two signal cores to transmit and detect logic levels, and is therefore able to tolerate significantly higher levels of common mode noise than RS-232, which uses the voltage between the signal core and a common earth. A brief summary of some typical characteristics of the two standards is given below.
Signal detection Differential Single-ended
Receiver threshold 200 mV +1.5 V
Meter output swing 0 to +5 V (no load) +2 to +3 V (120 ohm load)
A converter is required for communication between the two standards. Further details are given in Section 3.8.2.
Only two signal connections are required for RS-485, usually called A and B, sometimes ‘+’ and ‘–‘.
For areas which may experience high common mode signals, a third conductor can be used as a ground reference for the communications signals. If used, this should be connected to Terminal 2 (Power supply negative) on the meter.
Converters are available from a number of sources, and can range from simple in-line devices that simply plug into a PC’s RS-232 port, to programmable devices with full isolation between the two networks.
Note: The meter uses a half-duplex implementation of RS-485, such that the A and B signals are used for data transmission in both directions. This requires that the RTS line is toggled to indicate the transmission direction. This can be done by the host computer, or automatically by an RS-485/232 converter which has the facility to do so. If you are using Windows NT, 2000 or XP on your PC, you should use a converter which automatically changes RTS (as detailed below) otherwise the link may not work correctly.
For simple installations, where the following conditions are valid, a simple in-line converter will be satisfactory:

Micro Motion recommends the K2–ADE (Terminal Block type or DIN connector type) converter, manufactured by KK Systems Ltd that will work with Windows 98, NT, 2000 and XP. This converter is available through Micro Motion when you purchase the ADView software. The ADView software package includes the latest Windows version of the software, plus a K2-ADE RS-485/RS-232 converter (Terminal Block version).
Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction
The K2-ADE converter derives its power from the PC’s RS-232 port RTS or DTR line, which must be held permanently in the high state. This is normally adequate for short distances where there are only a few devices on the network. However, the ability of the port to supply sufficient power is not guaranteed, especially for laptop PCs, and it may be necessary to connect an external power supply. This may also be necessary if using Windows NT, 2000 or XP.
To check the voltage levels, measure the voltages on the RTS input (pin 7) and the DTR input (pin 4) while the converter is connected to the PC (or other RS-232 device). This procedure needs a break-out box (not supplied). Whichever input is powering the converter must have at least +6 V during communications. Where the power is found to be insufficient, a 9 VDC supply can be connected between Pin 9 (+) and Pin 5 (GND) of the RS-232 connector. Connections are shown in Figure 3-8. See also the manufacturer’s technical information for details.
For permanent installations and cables distances greater than 100 m
For permanent installations, and where the network length is more than 100 m or so, Micro Motion can supply the following DIN-rail mounted device from KK Systems Ltd.
The KD485–ADE is three-way isolated, providing isolation between the two ports and the power supply. It requires a +7 to +35 V power supply and typically takes 1 to 2 W; (power consumption is largely independent of supply voltage). It is capable of working with Windows 98, NT, 2000 and XP. For a PC running Windows NT/2000/XP, the RTS connection can be omitted.

Figure 3-9 Modbus connections > 100 m

The default configuration of the KD485-ADE has Port 2 configured for 9600 baud. This is the correct baud rate for the meter. (See Section 3.8.4 for details).
RS232 to RS485 Interface Converter/Isolator
Port 2
RS232
RS485
The switch on the KD485-ADE should be set with SW1 On (to enable half-duplex operation on Port 2), with the other three switches (SW2, SW3, SW4) set to Off.
When two or more devices are connected on the same RS-485 network, this is known as a multi-drop configuration (see Section 3.8.3). Each device must be configured with its unique slave address before being installed on the network.

3.8.3 RS-485 multi-drop
When several devices are connected in parallel on an RS-485 network, this is known as a multi-drop network. Although it is theoretically possible to have up to 256 devices, in practice this is limited to around 32 or less, depending largely on the driving power of the Master. Each device has a unique slave address. For the meter, this address must be individually programmed using the ADView or ProLink II (v2.9 or later) software, before being connected to the multi-drop network (see section 4.4.3 for details).

1 2 3 4 5
Wiring is quite straightforward: simply connect ‘B’ terminal to ‘A’ terminal, A to B. On some devices, the RS-485 signals may be marked + and –. The + signal generally corresponds to the A signal, and the – signal to B. If you encounter communication difficulties with RS-485, the first remedy is to swap over the ‘A’ and ‘B’ connections at one end of the network.
The meter’s RS-485 interface uses the following parameter settings, which are not selectable:
Port 2 RS485


Installation Using ADView and ProLink II Electrical Connections Introduction
4.1.1 What is ADView?
ADView is a software package provided by Micro Motion to enable you to:
ADView is installed on a PC and interacts with the density/viscosity meter through one of the PC’s standard serial (RS-232) ports.
ADView requires Microsoft’s Windows operating system: Windows 3.1, 95, 98, NT, 2000 or XP.
ADView provides many useful facilities, such as:
ADView software is available for the PC on a variety of media (for example, CD-ROM) and is freely available to download from the Micro Motion web site (at www.micromotion.com).

Start the ADView software by navigating through the Start Menu to the program entry of ADView 6. Left-click on it once and the window shown below will then appear.

Each of the six icons gives you access to the various facilities of ADView. You can choose to connect a Modbus device to one of the PC’s serial ports, or you can use ADView’s built-in simulation of the meter.

To run the simulation, choose Options > Simulate board response from the menu bar and choose the appropriate densitometer option. Then, click on the OK buttons, as necessary, to return to the main ADView screen. When simulation is chosen, ADView ignores the serial port and supplies simulated data. However, you do still need to click on the Communications Setup button followed by the Connect button. Then, click on the OK buttons, as necessary, to return to the main ADView screen.
Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction
To operate with a real Modbus device, you will need to connect it to a suitable power supply (see the technical manual for the device) and need a connection to a serial port on the PC. Full details for connecting to the Modbus (RS-485) link on the meter are in Chapter 4.
ADView automatically configures the selected port with the correct settings for the device. For the meter, this is 9600 baud rate, 8 data bits, no parity, 1 stop bit, and Xon/Xoff (software) flow control.
An interesting feature of Windows NT is that it does not allow the RTS line to be toggled directly; any attempt to do so will result in a crash or other problem. Unfortunately, some RS-485/232 converters require RTS to be toggled. To overcome this difficulty, ADView reads the OS environment variable to determine whether the operating system is Windows NT. If it is, ADView does not toggle RTS, and you will need to use an RS-232/485 adapter which automatically switches the data direction without using RTS.
To set the OS variable, click on the Start button, then choose Settings > Control Panel. Click on the System icon, and select the Environment tab. A list of environment variables and their values is shown. If OS does not appear in the list, type ‘OS’ (no speech marks) in the Variable text box, and ‘Windows_NT’ (no speech marks or spaces) in the Value box.
To check whether the link is working, you can use the auto-detect facility in ADView. Select the correct PC port, and then click on the Connect button in the Communications dialog box. ADView will set the port communications parameters, and then attempt to establish contact with any Modbus devices connected to the serial link, within the address limits set in the dialog box.


When it finds a device, the message box below appears:

If no active device is found, a warning message is given:

In this case, check that the device is powered up correctly, that the cables and adapter are pushed fully home, and that the communications settings on the device and selected serial port are the same.
ADView facilities
The main ADView window gives access to the various facilities available. A brief description of each is listed below. Using the facilities is largely intuitive so that you can quickly learn the system.
Communications Setup Sets up and checks RS-232/RS-485 communications.
Register dump/load With this facility you can dump the contents of all (or selected) Modbus registers from the device, or alternatively transmit values to them. File format is selectable (Excel/tab delimited, or Notepad/Space delimited).
Board Configuration · Data logging

Meter details Shows a list of meter details such as type, serial number, calibration dates, software version, etc.
Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction
Diagnostics Enables you to view: - live sensor readings - the status of the meter - values of working coefficients You can also verify calculations.
File Exit Exit ADView program.
Tools Health Check Determines whether the system is functioning correctly.
Register Read/Write A facility for reading or writing to any of the Modbus registers (see Section )
Direct Comms. Enables you to specify exactly what will be transmitted on the Serial link (see Appendix D).
Engineer Status Only used by Micro Motion service engineers.
Simulate board response/ Actual Board
Allows you to select between these two options
Enable / disable screensaver Allows you to select between these two options. When enabled, the screensaver operates as configured by the Windows system settings.
Provides a means of opening or selecting ADView’s facilities.
Help About ADView Displays software version number.
The factory configuration sets the slave address to 1. However, in many applications it will be necessary to allocate another address. In a multi-drop application, where several Modbus devices are connected on the same network, it is essential to configure unique slave addresses for each device.
To do this, you will need to run ADView and use the Register Read/Write facility, detailed in “Register Read / Write” on page 44. Check the value in Register 30 (Modbus Slave Address). If it is not the required value, enter the desired value and click on the write button. The meter will now be configured with the new slave address.
The board configuration controls the way in which the meter will process and present data, user settings, calibration constants and other factors. This data is stored in non-volatile memory known as registers; a full list of the registers used in the meter is given in Appendix D.
To configure the meter, it is necessary to write data into the configuration registers using the RS485/Modbus link. ADView provides a convenient and graphical way of doing this without you needing to know about register addresses and data formats.

Certain parameters are not available for configuration by ADView, including the Density Offset value which may be required to fine tune the calibration of the meter. However, ADView does have tools for reading and writing to individual Modbus registers (using the Tools > Register Read/Write facility), and for direct communication on the Modbus (using Tools > Direct Comms). More details and examples are given in Appendix D, but for the significant majority of applications these tools will not be required.
There is no facility within ADView or the meter to ‘reset’ to a default configuration. Therefore, before attempting any alterations to the configuration, you are strongly advised to use the Register Dump/Load facility in ADView to store the existing configuration (see “Register Dump / Load” on page 43). Then, if any mishap occurs, you will be able to restore the configuration from the saved file.
ADView’s Board Configuration window is shown below:

To exit from any of the configuration windows without making any changes, press the Esc key on your computer keyboard.
To configure the density referral calculation, you will need to enter the relevant information.
ADView’s Data Logging function is a useful tool for checking setups and performing experimental data capture. The diagram below explains some of the features.

Graphical representation of analog output.
For selecting the parameter to be
Select analog output of another transmitter.
Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction

Tabular display of instantaneous output of transmitter.
For multi-drop configurations, the output of up to three transmitters can be displayed simultaneously.
Click Start to start logging.
Click Stop to stop logging
Click OK to close Data Logging window
The Log Setup button – which is activated when logging has been stopped – enables you to configure the frequency of logging, where the logged data will be filed, and the format of the data.
Display Selection dropdown list to select the transmitter and parameter to be displayed on the graph
Click Show Graph to configure and display graph
This facility is essential for saving the configuration of your meter. You should use it to save the current configuration before you start to alter it, in order to restore it if things go wrong for any reason. Also, if you send the meter away for servicing or re-calibration, you should save the current configuration. Details are given below.

Enter desired filename for Dump, or required filename for Load.
Address of unit being accessed

Choose data delimiter (Dump only)
Choose which sets of registers to save to file, or simply save all of them.
You can also specify individual registers.
Restore a previously saved set of register data from file.
Store the selected register data to a file.
In a few cases, it may be useful to write directly to a single Modbus register. Two likely occasions for using this feature are to set the Slave Address of the unit and to configure a density offset. Appendix B has a complete list of the registers.
Before making any changes to individual registers, you should save the current configuration to a file to safeguard your configuration if anything goes wrong. See “Data Logging” for more information.
From ADView’s menu bar, select Tools > Register Read/Write.

To see a complete list of Modbus register numbers and descriptors, click here.
The current register number appears here.

Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction Installation Using ADView and ProLink II Electrical Connections Introduction
Choose the one you want to access.
For non-numerical values, click here to see complete list of possible entries and select one to write into the register.
The Read button causes the current value of the chosen register to be displayed.
Enter numerical values directly. The Write button causes the current value to be written to the selected register.
You can read and write to any number of registers. When you have done all you want to, click OK.
4.2.1 Overview
ProLink II is a Windows-based configuration and management tool for Micro Motion meters. It provides complete access to meter functions and data.
This chapter provides basic information for connecting ProLink II to your meter. The following topics and procedures are discussed:
The instructions in this manual assume that users are already familiar with ProLink II software. For more information on using ProLink II, see the ProLink II manual.
To use ProLink II with a 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter, the following are required:
- For RS-485 connections, an RS-485 to RS-232 signal converter.
- 25-pin to 9-pin adapter (if required by your PC)
Table 1-1 describes the options for connecting ProLink II to your meter.

Connection Physical layer Protocol
RS-485 terminals or RS-485 network RS-485 Modbus
ProLink II provides a configuration upload/download function which allows you to save configuration sets to your PC. This allows:
Micro Motion recommends that all meter configurations be downloaded to a PC as soon as the configuration is complete.
To access the configuration upload/download function:
ProLink II can be configured for the following languages:
To configure the ProLink II language, choose Tools > Options.
In this manual, English is used as the ProLink II language.

General Maintenance Calibration Check
5.1.1 Factory calibration
Prior to leaving the factory, the 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter is calibrated within a standard physical boundary (typically 52.5 mm diameter) against Transfer Standard instruments traceable to National Standards.
Three fluids ranging in density from 1 to 1000 kg/m3 are used to establish the general density equation constants K0, K 1, and K2 (see Section 5.1.5). The temperature coefficients (K18 and K19) are derived from the air-point and material properties.
The calibration procedure relies on units being immersed in fluids whose density is defined by Transfer Standards. Great attention is paid to producing temperature equilibrium between the fluid, the unit under test and the Transfer Standard (see Section 5.1.2). In this way, accurate calibration coefficients covering the required density range can be produced.
Viscosity calibration is achieved using three fluids (or four for the 0 – 10 cP range) with different calibrated viscosity values to derive the three general viscosity equation coefficients V0, V1, and V2 (see Section 5.1.4.)
All instruments are over-checked on water to verify the density calibration, and with two different fluids to check the viscosity calibration. This check is monitored by the Micro Motion Quality Assurance Department.
The Transfer Standards for viscosity calibration are fluids which have been accurately measured within the Micro Motion Standards Laboratory. For density calibration, Transfer Standard instruments used in the calibration are selected instruments which are calibrated by the British Calibration Service Calibration Laboratory and are certified.
Transfer Standard calibration uses a number of density-certified liquids, one of which is water. The densities of these reference liquids are obtained using the Primary Measurement System whereby glass sinkers of defined volume are weighed in samples of the liquids.
Calibration of the Transfer Standard instruments is performed under closely controlled laboratory conditions and a calibration certificate is issued. Calibrations are repeated, typically every six months, producing a well-documented density standard.

5.1.3 Instrument calibration
Each meter is issued with its own calibration which is programmed into the instrument electronics before it leaves the factory. Under normal circumstances it should not be necessary to re-calibrate the meter provided it is used in the environment for which it was calibrated originally.
The calibration data is shown on a calibration certificate supplied with the instrument. The calibration contains the following:
The values for all the V and K coefficients shown on the calibration certificate are programmed into the meter’s registers, and should not be altered.
The General Viscosity Equation, used to calibrate the meter and shown in the Calibration certificate is:
η = V0 + V1 / Q2 + V2 / Q4
where η is the calculated viscosity, Q is the quality factor of the tuning fork, and V0, V1 and V2 are viscosity coefficients, derived from the factory calibration data and selected to optimise the accuracy of the viscosity measurement across the calibrated viscosity range for the known physical conditions.
The General Density Equation, used to calibrate the meter and shown in the Calibration certificate is:
ρ = K0 + K1τ + K2τ2
where ρ is the calculated density, τ is the time period (in μs) of the tuning fork, and K0, K1 and K2 are density coefficients, derived from the factory calibration data and selected to optimise the accuracy of the density measurement across the calibrated density range.
Temperature effects are also compensated for using a second equation:
ρ’ = ρ( 1 + K18(t - 20)) + K19(t - 20)
where ρ’ is the new (temperature compensated) density value, t is the measurement temperature, and K18 and K19 are temperature correction coefficients.

General Maintenance Calibration Check
5.2.1 Viscomaster sample calibration certificate
CALIBRATION CERTIFICATE
VISCOSITY CALIBRATION AT 20ºC (2” Schedule 40)
VISCOSITY QUALITY VISCOSITY = V0 + V1.1/Q**2 + V2.1/Q**4 (cP) FACTOR INSTRUMENT CHECK DATA 0.59 920.79 10.67 268.64 AIR POINT (20ºC) QUALITY FACTOR = 5208 53.36 123.35 108.16 84.25
ULTRA-LOW RANGE LOW RANGE (0.59 – 10.67) (10.67 – 108.16)
V0 = -4.90450E-01 -1.40975E+00 V1 = 8.49042E+05 8.81886E+05 V2 = -3.13525E+09 -7.39203E+08
DENSITY CALIBRATION AT 20ºC (2” Schedule 40)
DENSITY TIME PERIOD B DENSITY = K0 + K1.TB + K2.TB**2 (kg/M3) (usec) K0 = -2.55203E+03 0 526.090 K1 = 1.33903E-02 (Air 525.982) K2 = 9.19530E-03 300 556.194 500 575.391 800 603.042 Dt = D ( 1 + K18(t-20) ) + K19(t-20) 1000 620.793 1600 671.239 K18 = -4.614E-04 K19 = -1.101E+00
Where D = Density (uncorrected) | FINAL TEST & | Dt = Density (temperature corrected) | INSPECTION | Dv = Density (temp and viscosity corrected) | | TB = Time Period B(uS) | | Q = Quality factor | | t = temperature (ºC) | | Ref No:- XXXXXX/Vx.x DATE : 16DEC04

5.2.2 Viscomaster Dynamic sample calibration certificate
CALIBRATION CERTIFICATE 7829FEANAJRBBA VISCOMETER SERIAL NO : XXXXXX CAL DATE : 03FEB05 PRESSURE TEST : 160 Bar
VISCOSITY CALIBRATION AT 20ºC (2” Schedule 40)
VISCOSITY QUALITY VISCOSITY = V0 + V1.1/Q**2 + V2.1/Q**4 (cP) FACTOR INSTRUMENT CHECK DATA 0.77 84.85 10.78 265.92 AIR POINT (20ºC) QUALITY FACTOR = 5094 53.30 123.35 107.75 85.60
ULTRA-LOW RANGE LOW RANGE (0.77 – 10.78) (10.78 – 107.75)
V0 = -1.18001E+00 -1.29351E+00 V1 = 9.43728E+05 8.59976E+05 V2 = -6.69315E+09 -4.46850E+08
Where TB = Time Period B(uS) | FINAL TEST & | Q = Quality factor | INSPECTION | t = temperature (ºC) | Ref No:- XXXXXX/Vx.x DATE : 03FEB05

General Maintenance Calibration Check
5.3.1 Ambient air calibration check
An air check is a simple and convenient method to see if any long term drift or corrosion and deposition on the tines has occurred.
An on-line calibration adjustment may be required if:
The meter is a very accurate and stable instrument, and will normally provide good measurements. If it is suspected of giving incorrect results, you should confirm this by carefully checking the integrity of the fluid temperature measurement, and compare this with the temperature measurement given by meter. You should also verify the integrity of the density check measurement. It is only after you have eliminated all other possible causes of error that you should attempt to make adjustments to the calibration of meter.
Normally the density calibration adjustment is made by configuring a simple density offset into the instrument. If a more detailed calibration adjustment is required, such as a two- or three-fluid calibration adjustment for offset and scale, then refer to Micro Motion.
For further details on these procedures, reference should be made to:

Energy Institute: HM7. Density, sediment and water. Section 1: General guidance on test methods (formerly PMM Part VII, S1)
1st ed 1996 ISBN 978-0-85293-154-7
Energy Institute: HM8. Density, sediment and water. Section 2: Continuous density measurement (formerly PMM Part VII, S2)
2nd ed Sept 1997 ISBN 978-0-85293-175-2
Chapter 14 - Natural Gas Fluids - Section 6: Installing and proving density meters used to measure hydrocarbon liquid with densities between 0.3 and 0.7 gm/cc at 15.56°C (60°F) and saturation vapour pressure, April 1991.

General Maintenance Calibration Check
Care is essential in handling of the meter during its removal from and fitment to the pipeline/tank and during transportation. Wherever possible, retain and use the original packaging.
The 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter is rugged and robust, and has no moving parts. When correctly installed and operated, servicing is not normally required, even with poor quality fluid, and no periodic maintenance procedure is specified. It is recommended that a visual inspection is carried out at intervals to check for leaks and physical damage, and corrective maintenance carried out when required.
ADView’s Data Logging facility can be used whenever necessary to verify that the meter is functioning correctly.
Check calibrations should be carried out at specified intervals in order to identify a malfunction or deterioration in meter performance. If a fault or a drop in performance is detected, further tests are required to identify the cause of the fault. Remedial action is limited to cleaning the meter tines, making good any poor connections, and replacing the internal electronics. In the extreme cases the complete meter may need to be replaced.
No periodic maintenance procedure is specified, but the following procedure is recommended for periodic inspection. It can also be used when fault finding.

6.2.2 Electrical check
If the current consumption is outside this range, contact Micro Motion.
When several systems are run in parallel and use the same fluid source, comparison of the line viscosity, base density and temperature readings between installations can be a useful indicator of possible system faults. Differences between readings, or changes from the normally observed conditions should always be investigated to confirm that instrumentation is functioning correctly.
A fault may be categorized as either an erratic reading or a reading which is outside limits.
Electrical faults can also cause symptoms which appear to affect the readings and it is recommended that the electrical system is checked first, before removing the meter for servicing.
Table 6-1 Faults and possible causes
Fault Possible causes Remedy
Readings fluctuate slightly, i.e., are noisy
Analog output averaging time not long enough
Increase the averaging time using ADView’s Board Configuration facility (see the Using ADView and ProLink II chapter).
Erratic readings One or more of: Gas bubbles around tines; cavitations; severe vibration or electrical interference; large amount of contaminants
Remove primary cause; e.g.: -install air release units to release gas; -apply back pressure to discourage formation of bubbles; -remove cause of vibration Alternatively, it may be necessary to adjust the Time Period Trap.
Readings outside limits Deposition and/or corrosion on the tines.
Clean tines.

Table 6-1 Faults and possible causes continued
General Maintenance Calibration Check
Analog output = 0 mA No power to analog output If voltage across pins 5 and 6 is not 15 to 28 V, replace power supply.
Analog output circuit failure Use ADView’s facility to set the analog output to 4, 12 or 20 mA (in Board Configuration) to check whether the output is functioning. If not, replace circuit boards.
Analog output is 2 mA Alarm condition caused by lack of power to meter
If voltage across pins 1 and 2 is not 20 to 28 V, check and replace main power supply.
Alarm condition caused by other internal failure
Use ADView Diagnostics to check that phase locked loop is in lock.
If analog output and Modbus appear to be functioning correctly, the temperature sensor has probably failed.
Return the meter to Micro Motion for servicing.
Viscosity reads high during normal running
Flow rate too low Increase flow or change to smaller flow-through chamber
Insulation defective Repair or replace insulation
PFA laminate damaged, leading to coating of fork tines
Remove meter for visual check; return to Micro Motion for servicing.
Calibration data is corrupted Compare calibration data to certificate or stored configuration. Reprogram as necessary.
Viscosity reads high after engine shutdown or restart
Pump coated with aspaltenes. Check pump delivery; service pump.
Bypass not fully closed. Close bypass.
PFA laminate damaged, leading to coating of fork tines
Remove meter for visual check; return to Micro Motion for servicing.
Calibration data is corrupted Compare calibration data to certificate or stored configuration. Reprogram as necessary; return to Micro Motion for servicing.
meter does not communicate with ADView
Power failure to meter Check power supply to meter and converter; replace if necessary
Power supply to RS-485/232 converter failed.
A and B Modbus connections reversed
RS-485/232 converter failed, wired incorrectly, or connected the wrong way round
ADView incorrectly installed on PC
Incorrect Slave address chosen for meter
RS-232 port on PC failed. Connect to another free RS-232 port on the PC, if available.
Alternatively connect a known working RS-232 device to the PC to check that the port is working.

6.3.2 Mechanical servicing
This mainly comprises the cleaning of any deposition or corrosion from the tines. Deposition is removed by the use of a suitable solvent. For corrosion, solvent and the careful use of a fine abrasive will usually be sufficient. Take care not to damage the PFA lamination if installed. However where extensive corrosion has been treated, it is highly recommended that a full calibration is carried out to check the meter characteristics.
Care is essential in handling the meter during transit, installation, and removal from the pipeline/tank.
Disturbances in the fluid caused by bubbles, cavitations or contaminants can cause sudden changes in the measured output, which may, under some circumstances, give rise to instability (i.e. hunting) in a control system relying on the measurement. The meter can maintain the analog output during such perturbations by ignoring the aberrant measurement, and maintaining the output at the last good measured value. This facility is known as the Time Period Trap (TPT).
Under all normal circumstances, the factory settings for the TPT should be used. However, in extreme cases it may be necessary to alter the settings to meet the demands of a particular system. This should only be done after monitoring the behavior of the system for some time, to establish the normal running conditions.
Great care must be taken not to reduce the sensitivity of the meter so that normal response to fluctuations in the fluid is impaired.
The time period trap facility works as follows:
After each measurement of the time period (of the meter’s vibrating tines) the new value is compared with the previous value. If the difference between them is smaller than the allowable tolerance, the output is updated to correspond to the new measured value, and the TPT remains inoperative; i.e., operation is normal. If the difference exceeds the allowable tolerance, the output remains at the its previous level, and does not follow the apparent sudden change in value.
This process is repeated until either of the following:
Two Modbus Registers control the operation of the Time Period Trap facility. These can be changed, if necessary, using ADView’s Register Read/Write facility.

General Maintenance Calibration Check


Calculated Parameters Modbus Communications Safety Certification Factory Default Settings
The 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter is supplied in a standard configuration. Analog Output 1 is set to provide line kinematic viscosity, in cSt. Analog Output 2 is set to provide line temperature, in °C. (Other units are kg/m3 for density; API base density derivation is also configured for the Viscomaster meter.)
The complete set of default values are shown below.
Viscomaster Factory default value
4 mA setting 0 0
20 mA setting As ordered As ordered
Units °C °C
4 mA setting 0 0
20 mA setting 150°C 150°C
General system Analog output User range
Low setting 0 0
High setting 100 100
ASTM D341 temperatures T1 and T2
ASTM D341 viscosities V1 and V2
0 cSt 0 cSt
Temperature offset 0 0
Pressure units bar bar
Pressure set value 1.013 1.013
Line density units kg/m3 kg/m3
Units cSt cSt · Hysteresis 2% 2% · Units cSt cSt

Parameter Viscomaster Dynamic Factory default value
Viscomaster Factory default value
Line density scale factor (Not Applicable) 1
Line density offset (Not Applicable) 0
User K0 (Not Applicable) +0000E+00
User K1 (Not Applicable) +0000E+00
Base temperature (Not Applicable) 15
Base pressure (Not Applicable) 1.013
Byte order Big Endian Big Endian
Register size 32 bit 32 bit

Calculated Parameters Modbus Communications Safety Certification Factory Default Settings
The 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter is capable of calculating a number of parameters based on the measured line dynamic viscosity, density and temperature. From these results, line kinematic viscosity can be calculated. The Viscomaster (not Viscomaster Dynamic) can also calculate parameters such as base (referred) kinematic viscosity, CII and CCAI.
Base density is the density of the fluid at a specified base (or referral) temperature which is different to the line (i.e., the actual) temperature of the fluid. Base density can be calculated by the API Referral method.
This calculation uses an iterative process to determine the density at the base temperature by applying temperature and pressure corrections using the API-ASTM-IP petroleum measurement tables.
The information required for the API density is:
Correction factors in the revised API-ASTM-IP petroleum measurement tables are based on the following correlation equations:
ρt / ρ15 = exp [-α15 Δt (1 + 0.8 α15 Δt)]
where:
The tangent coefficient differs for each of the major groups of hydrocarbons. It is obtained from the following relationship:
ρ + = α
ρ

where K0 and K1 are known as the API factors.
The hydrocarbon group can be selected as:
K0 and K1 are programmed into the meter for the first two groups. For refined products the values of K0 and K1 are automatically selected according to the corrected density:
Hydrocarbon Group
Gasolines 654 to 779 346.42278 0.43884
Jet Fuels 779 to 839 594.54180 0.0000
Fuel Oils 839 to 1075 186.9696 0.48618
For Crude Oil the API factors are:
Crude oil 613.972226 0.0000
User defined factors can be entered as any sensible value.
Isothermal secant compressibility can be defined by the simplified equation:
β = Isothermal secant compressibility at temperature t
δV1 = Change of volume from V0 to V1 P1 = Gauge pressure reading (P - 1.013) bars
0 P 1 β − = ρ ρ
ρ0 = Corrected density at zero (atmospheric) gauge.
ρ1 = Uncorrected density (Kg/m3)
P1 = (P - 1.013) where P is pressure in bars (P - base)

A correlation equation has been established for from the available compressibility data; such as,
Calculated Parameters Modbus Communications Safety Certification Factory Default Settings
loge C = -1.62080 + 0.00021592t + 0.87096 x 106(ρ15)-2 + 4.2092t x 103(ρ15)-2 per bar
Kinematic viscosity is defined as:
Base kinematic viscosity is the viscosity of the fluid at a specified base (or referral) temperature which is different to the line (i.e., the actual) temperature of the fluid. Base viscosity can be calculated using the kinematic–viscosity–temperature charts covered by the ASTM D341 standard.
The base kinematic viscosity of a petroleum oil or liquid hydrocarbon can be determined at any temperature within a limited range, if the kinematic viscosities at two temperatures are known.
The ASTM D341 standard charts use the following equation:
ν 0.7 + ( ) log log A B T log – =
The meter can be programmed with up to four ASTM curves that allow viscosity referral over a wide operating range. A ratio technique is employed when operating between the ASTM curves.
The ignition index (CII) is defined as:
CII = ( ) ( ) 7 . 0 v LOG LOG * 708 . 23 * 254565 . 0 t * 1038 . 0 795 . 270 15 + + ρ − +
where · where · where

Where:
The carbon aromaticity index (CCAI) is defined as:
CCAI = ( ) ( ) ( )⎟
⎛ + − + − − ρ 323 273 t LOG * 435 85 . 0 v LOG LOG * 141 81 15
Where:
⎞ ⎜ ⎝ · ⎠

Appendix C Safety Certification
Calculated Parameters Modbus Communications Safety Certification Factory Default Settings
Please contact Micro Motion if you need to have copies of the latest safety certification for the 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter.


Appendix D Modbus Communications
Calculated Parameters Modbus Communications Safety Certification Factory Default Settings Calculated Parameters Modbus Communications Safety Certification Factory Default Settings Calculated Parameters Modbus Communications Safety Certification Factory Default Settings Calculated Parameters Modbus Communications Safety Certification Factory Default Settings
The Modbus/RS-485 communications facility on the 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter can be useful in a number of ways. It is the only means of configuring the meter, and also gives access to diagnostic information not available on the analog output. Digital representations of the measured and calculated parameters are also available which lead to higher accuracy, and greater integration in digital networks and systems.
The RS-485 serial interface of the 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter communicates using the RTU Modbus protocol, which is a well established system used in many industrial applications. The protocol defines the way in which messages will be transmitted between Modbus devices, and details how the data will be formatted and ordered.
It is beyond the scope of this manual to give a full description of the protocol, but a useful reference on Modbus is the Modbus Protocol Reference Guide (PI-MBUS-200 Rev. D) (1992) published by Modicon Industrial Automation Systems Inc.
A Modbus network can have only one Master at any one time, with up to 32 Slaves. The 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter acts as a slave device, and only communicates on the network when it receives a request for information from a Master device such as a computer or a PLC.
The implementation used on the 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter is fully compliant with the Modicon Specification. All information is stored in memory locations in the 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter referred to as Modbus Registers. These store all the data required to control the operation, calculations and data output of the 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter. Modbus communication with the meter consists of reading or writing to these registers.
The 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter implements only two Modbus commands:-
Any number of registers can be read with Command 3, but only one register can be written to for each Command 16. This restriction does not limit the performance of the system, since all functions are mapped into the register structure in one way or another.
In most cases, it is unnecessary to understand the detail of the protocol, as this is taken care of by the application program. For example, the Micro Motion ADView or ProLink II software program enables you to configure the meter, and even read or write to individual Modbus registers, without you needing to know about Modbus.
However, if you are using a proprietary software package, or developing your own application software, the information given in this section will be invaluable.

D.2 Accessing Modbus registers
Any device which can drive the RS-485 interface on the 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter can, in theory, access the Modbus registers. In practice, some sort of user interface is required to simplify the process.
ADView offers several ways of accessing the registers.
Board Configuration: A graphical interface for viewing and setting the main configuration parameters of the 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter. Direct access to registers is not offered.
Register Read/Write This tool provides a simple window from which to read and write to named and numbered registers. When you write to a register, you are presented with a set of allowable values from which to choose. Thus the tool is only useful for communicating with Micro Motion meters. This is the simplest and most foolproof way of directly accessing the registers. Section 4 gives full details.
Direct Communications This is another tool which allows you to compose a sequence of data to be transmitted to/from the Modbus. This can be used to communicate with any Modbus device, providing that you know the register addresses, data format, indices, etc. The composition of the data is entirely up to the user, although the tool does compute and insert a checksum. Only those well versed in the use of Modbus protocol should attempt to use this facility. It is mainly designed for testing Modbus transmissions which are subsequently to be used in an application specific environment. A worked example of using this tool is given in section D.7.
If the meter Slave address or the values of Registers 47 and 48 are not known, Modbus communications cannot be carried out successfully, and it will be necessary to establish the current values in these items. If you are using ADView, you can search for the addresses of all connected slaves, and then interrogate the appropriate registers for each one.
If you are not using ADView, Section D.6 gives a procedure which will enable you to get this information.
D.3.1 Register size and content
All registers are 32 bits (whether they are integer or floating point types), although the Modbus specification states that registers are 16 bits and addresses and ‘number of register’ fields assume all registers are 16 bits long. All floating point values are in IEEE single precision format.
Registers are contiguous in the Modbus register ‘address space’. There is a one-to-one mapping of 32-bit meter register numbers to 16-bit Modbus register numbers. Therefore, only the full 32 bits of any register can be accessed. The upper and lower 16-bit segments have the same Modbus register number and consequently cannot be individually read.
Registers 47 and 48 within the meter allow the Modbus ‘dialect’ to be changed to suit the communicating device if it cannot easily be re-programmed. This is most easily done using ADView’s Register Read/Write tool (see the Using ADView and ProLink II chapter).
Their usage is as follows:

Modbus byte ordering
Calculated Parameters Modbus Communications Safety Certification Factory Default Settings Calculated Parameters Modbus Communications Safety Certification Factory Default Settings Calculated Parameters Modbus Communications Safety Certification Factory Default Settings Calculated Parameters Modbus Communications Safety Certification Factory Default Settings
0000000016 Big Endian (i.e. MSB first)
FFFFFFFF16 Little Endian (i.e. LSB first)
Register 48 contents Modbus register size
0000000016 16 bits
FFFFFFFF16 32 bits
In order to read 32-bit registers when Modbus registers are dealt with in units of 16 bits, you must specify twice the number of 32-bit register you want to read in the ‘number of registers’ field. For example, to read one 32-bit register, use '2'. If an attempt is made to read an odd number of registers, the command will fail.
In order to read 32-bit registers when Modbus registers are dealt with in units of 32 bits, you specify the actual number of registers you want in the ‘number of registers’ field. (for example, to read two 32-bit register in this mode, use '2'.
Each register is identified by a unique number, and the list is organized by this number. For each register, the contents are described, along with the data type of the contents.
The data type is always 32 bits unless stated otherwise. Variable names are given for reference purposes only. They have no other use.
In some cases the data in a register is used to represent a non-numerical quantity, known as an index. For example, the units of density can be kg/m3, gm/cc, lb/gal or lb/ft3 and these are represented by the numbers 91 to 94. Thus if Register 3 (line density) contains the value (index) 91, this means that the units of line density are kg/m3. Index values may, of course, be used for more than one register.
Tables of these indices are given in Section D.5
Index Table (where applicable)
Register Function Data Type
0 API product type Long integer D.5.1
1 API referral reference temperature 4-byte float
2 API referral reference pressure 4-byte float
3 Line density units Long integer D.5.2

Table D-1 Modbus register assignments continued
Register Function Data Type
4 Base density units Long integer D.5.2
5 Temperature units Long integer D.5.2
9 Output averaging time Long integer D.5.3
10 Analog Output 1 selected variable Long integer D.5.4
11 Analog Output 2 selected variable Long integer D.5.4
14 PWM factor for 4mA on Analog Output 1 Long integer
15 PWM factor for 20mA on Analog Output 1 Long integer
16 PWM factor for 4mA on Analog Output 2 Long integer
17 PWM factor for 20mA on Analog Output 2 Long integer
20 RTD calibration factor 4-byte float
21 Crystal oscillator calibration factor 4-byte float
22 Diagnostics flags Long integer
23 Line density value when fixed by diagnostics 4-byte float
24 Base density value when fixed by diagnostics 4-byte float
25 Temperature value when fixed by diagnostics 4-byte float
26 Pressure Units D.5.2
27 Referral temperature for matrix referral Long integer D.5.5
30 Modbus Slave address Long integer
47 Modbus byte order D.3.1
48 Modbus register size D.3.1
49 Software type Long integer D.5.6
53 2nd referral temperature for referred density 4-byte float
57 Full unit part number String
61 Hardware type Long integer D.5.7
64 Write-protected copy of RTD factor 4-byte float
65 Write-protected copy of crystal factor 4-byte float
66 Write-protected copy of Analog O/P 1 ‘4mA PWM factor’ Long integer
67 Write-protected copy of Analog O/P 1 ‘20mA PWM factor’ Long integer
68 Write-protected copy of Analog O/P 2 ‘4mA PWM factor’ Long integer
69 Write-protected copy of Analog O/P 2 ‘20mA PWM factor’ Long integer
72 Write-protected copy of Time Period Low Limit 4-byte float
73 Write-protected copy of Time Period High Limit 4-byte float
74 Write-protected copy of Q Factor Low Limit 4-byte float
75 Write-protected copy of Q Factor High Limit 4-byte float
127 Stored checksum for the FRAM Long integer
128 K0 4-byte float
129 K1 4-byte float
130 K2 4-byte float
131 K18 4-byte float

Table D-1 Modbus register assignments continued
Calculated Parameters Modbus Communications Safety Certification Factory Default Settings Calculated Parameters Modbus Communications Safety Certification Factory Default Settings Calculated Parameters Modbus Communications Safety Certification Factory Default Settings Calculated Parameters Modbus Communications Safety Certification Factory Default Settings
Register Function Data Type
132 K19 4-byte float
137 Meter time period trap count Long integer
138 Meter time period trap (difference in μs) 4-byte float
139 Time period value when fixed by diagnostics 4-byte float
140 Value represented by 4mA on analog output 4-byte float
141 Value represented by 20mA on analog output 4-byte float
146 Line pressure 4-byte float
172 Atmospheric pressure 4-byte float
173 Line density offset 4-byte float
174 Line density scaling factor 4-byte float
175 Special function calculation parameter A 4-byte float
176 Special function calculation parameter B 4-byte float
177 Special function calculation parameter C 4-byte float
178 Special function parameter d / density of water 4-byte float
179 Density of product A for special function calc. 4-byte float
180 Density of product B for special function calc. 4-byte float
181 Temperature offset 4-byte float
182 User K0 value for API referral 4-byte float
183 User K1 value for API referral 4-byte float
185 User range (alarm) high value 4-byte float
186 User range (alarm) low value 4-byte float
192 Write-protected copy of K0 4-byte float
193 Write-protected copy of K1 4-byte float
194 Write-protected copy of K2 4-byte float
195 Write-protected copy of K18 4-byte float
196 Write-protected copy of K19 4-byte float
201 Unit’s original calibration date Long integer
202 Unit’s most recent calibration date Long integer
203 Unit’s serial number Long integer
204 Unit type Long integer D.5.8
256 Status Register Long integer D.5.9
257 Corrected line density (1) 4-byte float
258 Corrected base density (1) 4-byte float
259 Line temperature (1) 4-byte float
260 Special function calculation result (1) 4-byte float
261 Meter time period (in μs) (1) 4-byte float
262 FRAM calculated checksums Long integer

Table D-1 Modbus register assignments continued
Register Function Data Type
263 RTD resistance (in ohms) (1) 4-byte float
264 Meter coil pickup level (in volts) (1) 4-byte float
265 Meter resonance Q value (2) 4-byte float
266 Electronics board temperature (in °C)
267/8 Software version string (1) String
286 Time period A (1) 4-byte float
287 Time period B (1) 4-byte float
288 Referred density at 2nd referral temperature (1) 4-byte float
289 Line dynamic viscosity (1) 4-byte float
290 Line kinematic viscosity (1) 4-byte float
294 Referred density API base density (15°C) (1) 4-byte float
320 Calculation selection (write-protected) Long integer
321 Phase angle fixing (write-protected) Long integer
322 Write-protected copy of V0 4-byte float
323 Write-protected copy of V1 4-byte float
324 Write-protected copy of V2 4-byte float
326 Write-protected copy of V3 4-byte float
327 Write-protected copy of V4 4-byte float
384 Value for Q when fixed 4-byte float
385 Value for time period B when fixed 4-byte float
386 Value for raw line density when fixed 4-byte float
387 Value for line dynamic viscosity when fixed 4-byte float
388 Value for line kinematic viscosity when fixed 4-byte float
400 Line dynamic viscosity scale factor 4-byte float
401 Line dynamic viscosity offset 4-byte float
402 Line kinematic viscosity scale factor 4-byte float
421 Line dynamic viscosity units Long integer D.5.10
422 V0 4-byte float
423 V1 4-byte float
424 V2 4-byte float
426 V3 4-byte float
427 V4 4-byte float
(1) This is a live value. Although it can be written to, it would be pointless. (2) This value is only valid when bit 3 (hex 08) is set in the diagnostics flag register (22), after a one-second pause.
This section provides an interpretation of the numerical indices used to represent non-numerical values.

D.5.1 API product type
Calculated Parameters Modbus Communications Safety Certification Factory Default Settings Calculated Parameters Modbus Communications Safety Certification Factory Default Settings Calculated Parameters Modbus Communications Safety Certification Factory Default Settings Calculated Parameters Modbus Communications Safety Certification Factory Default Settings
Used in Register 0. (The user values for K0 and K1 are stored in Registers 182 and 183.)
0 Crude (general crude)
1 Refined (general product)
2 User K0 and K1
Used in Registers 3, 4, 5 and 26.
10 kg / cm²
33 °F
90 SGU
91 g / cm³
92 kg / m³
93 lb / gal
94 lb / ft³
102 ° Baume heavy
D.5.3 Output averaging time
Used in Register 9.

D.5.4 Analog output selection
Used in Register 10.
9 Line dynamic viscosity
10 Line kinematic viscosity
11 Referred kinematic viscosity
Used in Register 27
0 Lowest temperature value in matrix
Highest temperature value in matrix
Used in Register 49.
1 API
D.5.7 Hardware type
Used in Registers 61.

D.5.8 Unit type
Calculated Parameters Modbus Communications Safety Certification Factory Default Settings Calculated Parameters Modbus Communications Safety Certification Factory Default Settings Calculated Parameters Modbus Communications Safety Certification Factory Default Settings Calculated Parameters Modbus Communications Safety Certification Factory Default Settings
Used on Register 204.
D.5.9 Status register flags
Used in Register 256.
0 00000001 ST_IN_LOCK P.L.L. is IN LOCK
1 00000002 ST_DIAG_ON DIAGnostics ON
2 00000004 ST_FT1_ALM 4 to 20 mA output 1 in ALarM
3 00000008 ST_FT2_ALM (1) 4 to 20 mA output 2 in ALarM
4 00000010 ST_FT3_ALM (1) 4 to 20 mA output 3 in ALarM
5 00000020 ST_HART_BOARD (1) whether HART BOARD is fitted
6 00000040 ST_RS232_BOARD (1) whether RS232 BOARD is fitted
7 00000080 ST_SWITCH_BOARD (1) whether SWITCH BOARD is fitted
8 00000100 ST_EXP0_BOARD (reserved for future expansion)
9 00000200 ST_EXP1_BOARD (reserved for future expansion)
10 00000400 ST_EXP2_BOARD (reserved for future expansion)
11 00000800 ST_EXP3_BOARD (reserved for future expansion)
12 00001000 ST_FT3_HART (1) HART is in control of its 4 to 20 mA output
13 00002000 ST_BAD_STATUS STATUS register corruption
14 00004000 ST_STAT_CORR one or more STATus registers have been CORRected
15 00008000 ST_TOTAL_DEATH status registers not updating - assume the worst
16 00010000 ST_USER_ALM User defined variable in alarm
21 00200000 ST_TEMP_HI TEMPerature reading too HIgh
22 00400000 ST_TEMP_LOW TEMPerature reading too LOW
23 00800000 ST_ROM_CSF ROM CheckSum Fail flag
24 01000000 ST_FRAM0_WPF FRAM0 Write Protect Fail
25 02000000 ST_FRAM1_WPF (1) FRAM1 Write Protect Fail
26 04000000 ST_FRAM0_RWE FRAM0 Read/Write Error
27 08000000 ST_FRAM1_RWE (1) FRAM1 Read/Write Error
28 10000000 ST_FRAM0_CSF FRAM0 CheckSum Fail flag

29 20000000 ST_FRAM1_CSF (1) FRAM1 CheckSum Fail flag
30 40000000 ST_FRAM0_ACK FRAM0 ACK/data error
31 80000000 ST_FRAM1_ACK (1) FRAM1 ACK/data error
(1) The status flags marked thus refer to hardware features not present in the meter. They can safely be ignored.
Used in Register 421.
1 P
D.6 Establishing Modbus communications
Using ADView, it is possible to establish which devices are available on the network, and their slave addresses. However, if you are not using ADView, the following procedure can be adopted.
If the meter Slave address or the values of Registers 47 and 48 are not known, Modbus communications cannot be carried out successfully, and it will be necessary to establish the current values in these items. The following procedure will do this.
The process is:
Make sure only the meter is connected to the Modbus Master, then send the following message (Read Register 47):
Register Address Checksum
00 03 00 4710 00 02 checksum
Wait for a response. If there is none, repeat the same message, with the Slave address changed to 1, and await a response. Repeat the process until a response is obtained. This will show the slave address of the meter.
Send the following message (Read Register 48), where nn is the meter slave address:
0 cP · 2 Pas · 3 mPas

Calculated Parameters Modbus Communications Safety Certification Factory Default Settings Calculated Parameters Modbus Communications Safety Certification Factory Default Settings Calculated Parameters Modbus Communications Safety Certification Factory Default Settings Calculated Parameters Modbus Communications Safety Certification Factory Default Settings
Register Address Checksum
nn 03 00 4810 00 02 checksum
The meter will respond with the following to show that the meter is set to 16-bits register size:
nn 03 04 4 data bytes checksum
Or, the meter will respond with the following to show that the meter is set to 32-bits register size.
nn 03 08 8 data bytes checksum
Thus, by reading the third byte of the response, you can deduce the value of Register 48.
Send the following message (Read Register 47), where nn is the meter’s slave address:
Register Address Checksum
nn 03 00 4710 00 02 checksum
The meter will respond with one of the following:
nn 03 04 4 data bytes checksum
nn 03 08 8 data bytes checksum
Examine the first four bytes of the data. If they are all 00, then the meter is in Big Endian mode; if they are all FF, then the mode is Little Endian.

D.7 Example of direct Modbus access
In many applications, direct access to Modbus will be unnecessary. ADView provides a way of configuring the 7829 Viscomaster® / Viscomaster Dynamic™ viscosity meter, and for accessing individual registers. This example describes how to access the meter directly, without the help of ADView.
However, before you start, you should configure the meter using ADView (described in the Using ADView and ProLink II chapter), and also set the Modbus Byte Order and Register Size (see Modbus Communications appendix).
The meter is assumed to have been configured with Register Size = 16-bit (Register 48 = 0), and has slave address = 1.
The following string will read the line density, which is held in Register 257 (010116).
Checksum (Automatically inserted if you are using ADView.) Slave address (hex)
Register address Hi byte
Register address Lo byte
Number of registers to read (Hi byte)
Number of registers to read (Lo byte)
The reply from the meter will be:
Checksum Slave address (hex)
Reply byte count

D.7.2 Example 2: Reading line density (32-bit register size)
Calculated Parameters Modbus Communications Safety Certification Factory Default Settings Calculated Parameters Modbus Communications Safety Certification Factory Default Settings Calculated Parameters Modbus Communications Safety Certification Factory Default Settings Calculated Parameters Modbus Communications Safety Certification Factory Default Settings
The meter is assumed to have been configured with Register Size = 32-bit (Register 48 = FFFF16), and has slave address = 1.
The following string will read the line density, which is held in Register 257 (010116).
Checksum (Automatically inserted if you are using ADView.) Slave address (hex)
Register address Hi byte
Register address Lo byte
Number of registers to read (Hi byte)
Number of registers to read (Lo byte
The reply from the meter will be the same as for Example 1.
Checksum Slave address (hex)
Reply byte count


Return Policy Product Data
E.1.1 Crude oil
Table E-1 Crude oil
Temp. (°C) Density (kg/m3) 60 738.91 765.06 791.94 817.15 843.11 869.01 894.86 920.87 946.46
55 742.96 768.98 794.93 820.83 846.68 872.48 898.24 923.95 949.63
50 747.00 772.89 798.72 824.51 850.25 875.94 901.80 927.23 952.82
45 751.03 776.79 802.50 828.17 853.81 879.40 904.96 930.50 956.00
40 755.05 780.68 806.27 831.83 857.36 882.85 908.32 933.76 959.18
35 759.06 784.57 810.04 835.48 860.90 886.30 911.67 937.02 962.36
30 763.06 788.44 813.79 839.12 864.44 889.73 915.01 940.28 965.53
25 767.05 792.30 817.54 842.76 867.97 893.16 918.35 943.52 968.89
20 771.03 796.18 821.27 846.38 871.49 896.59 921.68 946.77 971.85
15.556 774.56 799.57 824.59 849.60 874.61 899.62 924.63 949.64 974.65
15 775.00 800.00 825.00 850.00 875.00 900.00 925.00 950.00 975.00
10 778.95 803.83 828.72 853.61 878.50 903.41 928.32 953.23 978.15
5 782.90 807.65 832.42 857.20 882.00 906.81 931.62 958.45 981.29
0 786.83 811.46 836.12 860.79 885.49 910.21 934.92 959.66 984.42
Table E-2 Refined products
Temp. (°C) Density (kg/m 3 )
60 605.51 657.32 708.88 766.17 817.90 868.47 918.99 969.45 1019.87
55 610.59 662.12 713.50 769.97 821.49 872.00 922.46 972.87 1023.24
50 615.51 666.91 718.11 773.75 825.08 875.53 925.92 976.28 1026.60
45 620.49 671.68 722.71 777.53 828.67 879.04 929.38 979.69 1029.96
40 625.45 676.44 727.29 781.30 832.24 882.56 932.84 983.09 1033.32
35 630.40 681.18 731.86 785.86 835.81 886.06 938.28 986.48 1038.67
30 635.33 685.92 736.42 788.81 839.37 889.56 939.72 989.87 1040.01
25 640.24 690.63 740.96 792.55 842.92 893.04 943.16 993.26 1043.35
20 645.13 695.32 745.49 796.28 846.46 896.53 846.58 996.63 1046.68
15.556 649.46 699.48 749.50 799.59 849.61 899.61 949.62 999.63 1049.63

Table E-2 Refined products continued
15 650.00 700.00 750.00 800.00 850.00 900.00 950.00 1000.00 1050.00
10 654.85 704.66 754.50 803.71 853.53 903.47 953.41 1003.36 1053.32
5 659.67 709.30 758.97 807.41 857.04 906.92 956.81 1006.72 1056.63
0 664.47 713.92 763.44 811.10 860.55 910.37 960.20 1010.07 1059.93
The above tables are derived from equations, which form the basis of the data in the Revised Petroleum Measurement Tables (IP 200, ASTM D1250, API 2540 and ISO R91 Addendum 1).
The density temperature relationship used is:
( ) [ ] t 15 t 15 15
t 8 . 0 1 exp Δ α + Δ α − = ρ
= Density at base temperature 15°C (kg/m3)
= t°C –15°C (such as t – base temperature)
= Tangent thermal expansion coefficient per °C at base temperature 15°C
The tangent thermal expansion coefficient differs for each of the major groups of hydrocarbons. It is obtained using the following relationship:
ρ + = α
Crude Oil 771 – 981 613.97226 0.00000
Gasolines 654 – 779 346.42278 0.43884
Kerosines 779 – 839 594.54180 0.00000
Fuel Oils 839 – 1075 186.96960 0.48618
Table E-3 Platimum resistance law (To DIN 43 760)
°C Ohms °C Ohms °C Ohms °C Ohms °F Ohms °F Ohms
–50 80.31 5 101.91 60 123.24 115 144.17 0 93.03 100 114.68
–45 82.29 10 103.90 65 125.16 120 146.06 10 95.21 110 116.83
–40 84.27 15 105.85 70 127.07 125 147.94 20 97.39 120 118.97
–35 86.25 20 107.79 75 128.98 130 149.82 30 99.57 130 121.11
–30 88.22 25 109.73 80 130.89 135 151.70 32 100.00 140 123.24
–25 90.19 30 111.67 85 132.80 140 153.58 40 101.74 150 125.37
–20 92.16 35 113.61 90 134.70 145 155.45 50 103.90 160 127.50
–15 94.12 40 115.54 95 136.60 150 157.31 60 106.07 170 129.62
–10 96.09 45 117.47 100 138.50 155 159.18 70 108.23 180 131.74
–5 98.04 50 119.40 105 140.39 160 161.04 80 110.38 190 133.86
0 100.00 55 121.32 110 142.29 165 162.90 90 112.53 200 135.97

E.1.4 Density of ambient air
Taken at a relative humidity of 50%.
Return Policy Product Data
Air Pressure Air Temperature (°C)
(mb) 6 10 14 18 22 26 30
900 1.122 1.105 1.089 1.073 1.057 1.041 1.025
930 1.159 1.142 1.125 1.109 1.092 1.076 1.060
960 1.197 1.179 1.162 1.145 1.128 1.111 1.094
990 1.234 1.216 1.198 1.180 1.163 1.146 1.129
1020 1.271 1.253 1.234 1.216 1.199 1.181 1.163
Use pure, bubble-free water.
Table E-5 Density of water (in kg/m3 to ITS – 90 temperature scale)
Temp °C 0 2 4 6 8 10 12 14 16 18
0 999.840 999.940 999.972 999.940 999.848 999.699 999.497 999.244 998.943 998.595
20 998.203 997.769 997.295 996.782 996.231 995.645 995.024 994.369 993.681 992.962
40 992.212 991.432 990.623 989.786 988.922 988.030 987.113 986.169 985.201 984.208
60 983.191 982.150 981.086 980.000 978.890 977.759 976.607 975.432 974.237 973.021
80 971.785 970.528 969.252 967.955 966.640 965.305 963.950 962.577 961.185 959.774
100 958.345

E.1.6 Velocity of sound in liquids
Table E-6 Velocity of sound in liquids
Liquid Temp. (t °C) Velocity of Sound
Rate of Change
c t / c δ δ
( ms –1 K –1 )
Acetic acid 20 1173
Acetone 20 1190 –4.5
Amyl acetate 29 1173
Aniline 20 1656 –4.0
Benzene 20 1320 –5.0
Blood (horse) 37 1571
Butyl acetate 30 1172 –3.2
Carbon disulphide 25 1142
Carbon tetrachloride 20 940 –3.0
Chlorine 20 850 –3.8
Chlorobenzene 20 1290 –4.3
Chloroform 20 990 –3.3
Ethanol amide 25 1724 –3.4
Ethyl acetate 30 1133 –3.9
Ethyl alcohol 20 1162 –3.6
Formic acid 20 1360 –3.5
Heptane 20 1160 –4.5
n-Hexane 30 1060
Kerosene 25 1315 –3.6
Menthol 50 1271
Methyl acetate 30 1131 –3.7
Methyl alcohol 20 1121 –3.5
Methylene Chloride 25 1070
Nitrogen –189 745 –10.6
Nonane 20 1248
Oil (castor) 19 1500 –4.1
Oil (olive) 22 1440 –2.8
Octane 20 1197
Oxygen –186 950 –6.9

Table E-6 Velocity of sound in liquids continued
Rate of Change
Return Policy Product Data
( ms–1K–1) c t / c δ δ
n-Pentane 20 1044 –4.2
n-Propyl acetate 26 1182
Toluene 20 1320 –4.3
Turpentine 25 1225
Water (distilled) 10 1447.2
20 1482.3
30 1509.1
50 1542.5
70 1554.8
Water (sea) –4 1430.2
00 1449.5
05 1471.1
15 1507.1
25 1534.7
o-Xylene 22 1352


Return Policy Product Data
Micro Motion procedures must be followed when returning equipment. These procedures ensure legal compliance with government transportation agencies and help provide a safe working environment for Micro Motion employees. Failure to follow Micro Motion procedures will result in your equipment being refused delivery.
Information on return procedures and forms is available on our web support system at www.micromotion.com, or by phoning the Micro Motion Customer Service department.
Only equipment that has not been removed from the original shipping package will be considered new and unused. New and unused equipment requires a completed Return Materials Authorization form.
All equipment that is not classified as new and unused is considered used. This equipment must be completely decontaminated and cleaned before being returned.
Used equipment must be accompanied by a completed Return Materials Authorization form and a Decontamination Statement for all process fluids that have been in contact with the equipment. If a Decontamination Statement cannot be completed (for example, for food-grade process fluids), you must include a statement certifying decontamination and documenting all foreign substances that have come in contact with the equipment.




For the latest Micro Motion product specifications, view the PRODUCTS section of our web site at www.micromotion.com
Micro Motion USA Worldwide Headquarters 7070 Winchester Circle Boulder, CO 80301 T +1 303-527-5200 +1 303-522-6277 F +1 303-530-8459
Micro Motion Slough Emerson Process Management 158 Edinburgh Avenue Slough, Berks, UK SL1 4UE T +44 1753 756600 F +44 1753 823589
Micro Motion Asia Emerson Process Management 1 Pandan Crescent Singapore 128461 Republic of Singapore T +65 6777-8211 F +65 6770-8003
Micro Motion United Kingdom Emerson Process Management Limited Horsfield Way Bredbury Industrial Estate Stockport SK6 2SU U.K. T +44 0870 240 1978 F +44 0800 966 181
Micro Motion Japan Emerson Process Management 1-2-5, Higashi Shinagawa Shinagawa-ku Tokyo 140-0002 Japan T +81 3 5769-6803 F +81 3 5769-6844
Pages referenced in this manual — manufacturer, support and downloads.
Common problems, likely causes, and what to do.
| Problem | Possible cause | Remedy | |
|---|---|---|---|
| Readings fluctuate slightly, i.e., are noisy | Analog output averaging time not long enough | Increase the averaging time using ADView’s Board Configuration facility (see the Using ADView and ProLink II chapter). | p.60 |
| Erratic readings | One or more of: Gas bubbles around tines; cavitations; severe vibration or electrical interference; large amount of contaminants | Remove primary cause; e.g.: -install air release units to release gas; -apply back pressure to discourage formation of bubbles; -remove cause of vibration Alternatively, it may be necessary to adjust the Time Period Trap. | p.60 |
| Readings outside limits | Deposition and/or corrosion on the tines. | Clean tines. | p.60 |
| Analog output = 0 mA | No power to analog output | If voltage across pins 5 and 6 is not 15 to 28 V, replace power supply. | p.61 |
| Analog output = 0 mA | Analog output circuit failure | Use ADView’s facility to set the analog output to 4, 12 or 20 mA (in Board Configuration) to check whether the output is functioning. If not, replace circuit boards. | p.61 |
| Analog output is 2 mA | Alarm condition caused by lack of power to meter | If voltage across pins 1 and 2 is not 20 to 28 V, check and replace main power supply. | p.61 |
| Analog output is 2 mA | Alarm condition caused by other internal failure | Use ADView Diagnostics to check that phase locked loop is in lock. | p.61 |
| Temperature readings incorrect | If analog output and Modbus appear to be functioning correctly, the temperature sensor has probably failed. | Return the meter to Micro Motion for servicing. | p.61 |
| Viscosity reads high during normal running | Flow rate too low | Increase flow or change to smaller flow-through chamber | p.61 |
| Viscosity reads high during normal running | Insulation defective | Repair or replace insulation | p.61 |
| Viscosity reads high during normal running | PFA laminate damaged, leading to coating of fork tines | Remove meter for visual check; return to Micro Motion for servicing. | p.61 |
| Viscosity reads high during normal running | Calibration data is corrupted | Compare calibration data to certificate or stored configuration. Reprogram as necessary. | p.61 |
| Viscosity reads high after engine shutdown or restart | Pump coated with aspaltenes. | Check pump delivery; service pump. | p.61 |
| Viscosity reads high after engine shutdown or restart | Bypass not fully closed. | Close bypass. | p.61 |
| Viscosity reads high after engine shutdown or restart | PFA laminate damaged, leading to coating of fork tines | Remove meter for visual check; return to Micro Motion for servicing. | p.61 |
| Viscosity reads high after engine shutdown or restart | Calibration data is corrupted | Compare calibration data to certificate or stored configuration. Reprogram as necessary; return to Micro Motion for servicing. | p.61 |
| meter does not communicate with ADView | Power failure to meter | Check power supply to meter and converter; replace if necessary | p.61 |
| meter does not communicate with ADView | Power supply to RS-485/232 converter failed. | Check wiring | p.61 |
| meter does not communicate with ADView | A and B Modbus connections reversed | Check wiring | p.61 |
| meter does not communicate with ADView | RS-485/232 converter failed, wired incorrectly, or connected the wrong way round | Try another converter | p.61 |
| meter does not communicate with ADView | ADView incorrectly installed on PC | Re-install ADView | p.61 |
| meter does not communicate with ADView | Incorrect Slave address chosen for meter | Check slave address | p.61 |
| meter does not communicate with ADView | RS-232 port on PC failed. | Connect to another free RS-232 port on the PC, if available. Alternatively connect a known working RS-232 device to the PC to check that the port is working. | p.61 |