Instruction Sheet PN 51A-396/rev.F January 2011

Models 396 and 398

pH/ORP Sensors

For additional information, please visit our website

at www.emersonprocess.com/raihome/liquid/.

SPECIFICATIONS pH Range*AccuGlass: 0-14 pH Wetted Materials:

Minimum Conductivity: 75 μS/cm, nominal 100 μS/cm Weight/Shipping Weight: 0.45 kg/0.9 kg (1 lb/2 lb) Process Connections: None: must use 1 in. MNPT

process connector PN 23166-00 or 23166-01

Model 396polypropylene, 316 stainless steel,
glass, EP Model 398Tefzel; titanium; glass; and choice of

Kalrez, Viton or EPDM o-rings Interconnecting Cable: 15 ft (4.6 m), for use with

remote preamplifier Temperature Rating0 to 100° C (32 to 212° F) Pressure Range:
*PERCENT LINEARITY pH rangehemi bulb GPHT flat bulb GPLR 0-294%93% 2-1299%98% 12-1397%95% 13-1492%—

Model 396:

Hemi bulb: 100 - 1136 kPa (abs) (0 - 150 psig Flat bulb: 100 - 790 kPa (abs) (0 - 100 psig) Model 398:

Hemi bulb100 - 1825 kPa (abs) (0 - 250 psig)
Flat bulb100 - 790 kPa (abs) (0 - 100 psig)

CAUTION

WARNING

WARNING

SENSOR/PROCESS APPLICATION COMPATIBILITY

Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level!

The wetted sensor materials may not be compatible with process com position and operating conditions. Application compat ibility is entirely the responsi- bility of the user.

WARNING DANGER

Glass electrode must be wetted at all times (in storage and in line) to maximize sensor life.

Do not connect sensor to power lines. Serious injury may result.

ATEX DIRECTIVE

Special Conditions for safe use 1. All pH/ORP sensors have a plastic enclosure which must only be cleaned with a damp cloth to avoid the danger due to a build up of an electrostatic charge. 2. All pH/ORP sensor Models are intended to be in contact with the process fluid and may not meet the 500V r.m.s. a.c. test to earth. This must be taken into consideration at installation.

Page 2 of the Emerson Tupelo pH Sensor (396) user manual — NOTE
Page 2 · click to enlarge
Page 2

STORAGE

1. Immerse sensor in the first buffer solution. Allow sensor to equilibrate to the buffer temperature (to avoid errors due to temperature differences between the buffer solution and sensor temperature) and wait for reading to stabilize. Value of buffer can now be acknowledged by analyzer/transmitter.
1. It is recommended that electrodes be stored in their original shipping containers until needed.
2. Do not store at temperatures below -5°C (23°F).
3. Electrodes should be stored with a protective cap containing KCl solution (PN 9210342).
2. Once the first buffer has been acknowledged by the analyzer/transmitter, rinse the buffer solution off of the sensor with distilled or deionized water.
4. For overnight storage, immerse the sensor in tap water or 4 pH buffer solution.
3. Repeat steps 1 & 2 using the second buffer solution.
5. A pH glass electrode does have a limited shelf life of one year.
4. The theoretical slope value, according to the Nernst equation for calculating pH, is approximately 59.17 mV/pH. Over time the sensor will age, both in the process and in storage, and will result in reduced slope values. To ensure accurate readings, it is recommended that the electrode be replaced when the slope value falls below 47 to 49 mV/pH.

ELECTRODE PREPARATION

1. Remove electrode from shipping container.
2. Remove the protective boot covering the electrode bulb.
3. For pH electrode, rinse away salt film with clean water, then shake the electrode so that the internal solution fills the bulb, thus removing any air trapped there.

RECOMMENDED pH SENSOR STANDARDIZATION

For maximum accuracy, the sensor can be standard- ized on-line or with a process grab sample after a buffer calibration has been performed and the sensor has been conditioned to the process. Standardization accounts for the sensor junction potential and other interferences. Standardization will not change the sen- sor’s slope but will simply adjust the analyzer’s reading to match that of the known process pH.

CAUTION

The buffer in the protective boot may cause skin or eye irritation.

INSTALLATION

For sensor dimensions, see Figure 1.

MAINTENANCE FOR pH ELECTRODE

For sensor orientation, see Figures 2 and 3.

Electrodes should respond rapidly. Sluggishness, off- sets, and erratic readings are indicators that the elec- trodes may need cleaning or replacement.

For wiring, see Figures 4 - 18.

1. To remove oil deposit, clean the electrode with a mild non-abrasive detergent.

TWO POINT BUFFER CALIBRATION

2. To remove scale deposits, soak electrodes for 30 to 60 minutes in a 5% hydrochloric acid solution.

Select two stable buffer solutions, preferably pH 4.0 and 10.0 (pH buffers other than pH 4.0 and pH 10.0 can be used as long as the pH values are at least two pH units apart).

3. Temperature effect on life expectancy: If glass electrode life expectancy is 100% @ 25°C (77°F), then it will be approximately 25% @ 80°C (176°F).

NOTE

A pH 7 buffer solution reads a mV value of approx. zero, and pH buffers read approximately ± 59.1 mV for each pH unit above or below pH 7. Check the pH buffer manufacturer specifications for millivolt values at various temperatures since it may affect the actual value of the buffer solution mV/pH value.

Page 3 of the Emerson Tupelo pH Sensor (396) user manual — ORP CALIBRATION
Page 3 · click to enlarge
Page 3

ORP CALIBRATION

MAINTENANCE FOR ORP ELECTRODE

1. After making an electrical connection between the sensor and the instrument, obtain a standard solution of saturated quinhydrone. This can also be made quite simply by adding a few crystals of quinhydrone to either pH 4 or pH 7 buffer. Quinhydrone is only slightly soluble, therefore only a few crystals will be required.

Electrodes should respond rapidly. Sluggishness, off- sets, and erratic readings are indicators that the elec- trodes may need cleaning or replacement.

1. To remove oil deposit, clean the electrode with a mild non-abrasive detergent.
2. To remove scale deposits, soak electrodes for 30 to 60 minutes in a 5% hydrochloric acid solution.
2. Immerse the sensor in the standard solution. Allow 1-2 minutes for the ORP sensor to stabilize.
3. ORP (metallic) electrodes should be polished with moistened baking soda.

3. Adjust the standardize control of the instrument to the solution value shown in the table below. The resulting potentials, measured with a clean plat- inum electrode and saturated KCl/AgCl reference electrode, should be within +/- 20 millivolts of the value shown in the table below. Solution tempera- ture must be noted to ensure accurate interpreta- tion of results. The ORP value of saturated quinhy- drone solution is not stable over long periods of time. Therefore, these standards should be made up fresh each time they are used.

ORP of Saturated Quinhydrone Solution (millivolts) pH 4 SolutionpH 7 Solution Temp °C202530202530 mV Potential268264260948780
4. Remove the sensor from the buffer, rinse, and install in the process.
Page 4 of the Emerson Tupelo pH Sensor (396) user manual — INCH
Page 4 · click to enlarge
Page 4

MILLIMETER

Emerson Tupelo pH Sensor (396) — MILLIMETER diagram (page 4)
Fig. 1

INCH

FIGURE 1. Sensor Dimensions

Page 5 of the Emerson Tupelo pH Sensor (396) user manual — MILLIMETER
Page 5 · click to enlarge
Page 5
Emerson Tupelo pH Sensor (396) — MILLIMETER diagram (page 5)
Fig. 2

FIGURE 2. Flow Through and Insertion Installation For Models 396 and 398

1-1/2 inch Pipe Tee (PN 2002011) with 1 inch threaded connections.

Page 6 of the Emerson Tupelo pH Sensor (396) user manual — MILLIMETER
Page 6 · click to enlarge
Page 6
Emerson Tupelo pH Sensor (396) — MILLIMETER diagram (page 6)
Fig. 3

MILLIMETER

INCH

Emerson Tupelo pH Sensor (396) — MILLIMETER diagram (page 6)
Fig. 4
Emerson Tupelo pH Sensor (396) — MILLIMETER diagram (page 6)
Fig. 5

FIGURE 3. Submersion Installations For Models 396 and 398

Page 7 of the Emerson Tupelo pH Sensor (396) user manual — MILLIMETER
Page 7 · click to enlarge
Page 7
Emerson Tupelo pH Sensor (396) — MILLIMETER diagram (page 7)
Fig. 6
Emerson Tupelo pH Sensor (396) — MILLIMETER diagram (page 7)
Fig. 7

FIGURE 5. Models 396 and 398 Option -62 for Wiring to Models 1054, 1181, 2081, 54e, 81, 3081, 4081, 5081-P, Xmt,

FIGURE 4. Models 396 and 398 Standard Wiring

Option used with Models 1054A, 1054B, 2054,

SCL-P/Q, and 2700 Preamplifier (PN 23054-03). Also for

and Preamplifiers PN 22698-02 and 22698-03

use with remote Junction Box PN 23555-00.

Emerson Tupelo pH Sensor (396) — MILLIMETER diagram (page 7)
Fig. 8

Figure 6. Wiring Details for Model 396-54 (50) and 398-54 (50). For use with Junction Box PN 23309-04 (03) and Remote Preamplifier.

Page 8 of the Emerson Tupelo pH Sensor (396) user manual — SENSOR
Page 8 · click to enlarge
Page 8
Emerson Tupelo pH Sensor (396) — SENSOR diagram (page 8)
Fig. 9

MILLIMETER

INCH

MODELS 54, 54e, 3081, 81, 4081, 5081-P pH/ORP

SENSOR

FIGURE 7. Wiring to Junction Box (PN 23555-00) and Remote Preamplifier.

Page 9 of the Emerson Tupelo pH Sensor (396) user manual — SENSOR
Page 9 · click to enlarge
Page 9
Emerson Tupelo pH Sensor (396) — SENSOR diagram (page 9)
Fig. 10
Emerson Tupelo pH Sensor (396) — SENSOR diagram (page 9)
Fig. 11

FIGURE 9. Wiring to Models 3081, 4081,

FIGURE 8. Wiring to Model Xmt

Emerson Tupelo pH Sensor (396) — SENSOR diagram (page 9)
Fig. 12

FIGURE 10. Wiring to Model 1055-22-32.

and 5081-P

Page 10 of the Emerson Tupelo pH Sensor (396) user manual — SENSOR
Page 10 · click to enlarge
Page 10
Emerson Tupelo pH Sensor (396) — SENSOR diagram (page 10)
Fig. 13

FIGURE 11. Wiring to Model 1055-22-32 through a Remote Junction Box

Emerson Tupelo pH Sensor (396) — SENSOR diagram (page 10)
Fig. 14

FIGURE 12. Wiring to Model 1055-01-10-22-32

Page 11 of the Emerson Tupelo pH Sensor (396) user manual — SENSOR
Page 11 · click to enlarge
Page 11
Emerson Tupelo pH Sensor (396) — SENSOR diagram (page 11)
Fig. 15

FIGURE 13. Preparing Coax Cable with BNC Connector for Wiring to

Models 54, 54e, 81, 2081, 3081, 4081, 5081-P, and Xmt-P

Emerson Tupelo pH Sensor (396) — SENSOR diagram (page 11)
Fig. 16

FIGURE 14. Preparing Coax Cable for Wiring to Model 1181

Page 12 of the Emerson Tupelo pH Sensor (396) user manual — SENSOR
Page 12 · click to enlarge
Page 12
Emerson Tupelo pH Sensor (396) — SENSOR diagram (page 12)
Fig. 17
Emerson Tupelo pH Sensor (396) — SENSOR diagram (page 12)
Fig. 18

For more detailed instructions, see electronic file 51-396to398 on “Instruction Manuals” CD-rom.

Emerson Tupelo pH Sensor (396) — SENSOR diagram (page 12)
Fig. 19

The right people, the right answers, right now. ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE

Emerson Tupelo pH Sensor (396) — SENSOR diagram (page 12)
Fig. 20

2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250

http://www.raihome.com

© Rosemount Analytical Inc. 2011

Emerson Process Management

Sources

References & links

Pages referenced in this manual — manufacturer, support and downloads.

Ask about the Emerson Tupelo pH Sensor (396)

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Emerson Tupelo pH Sensor (396)

Be the first to ask — other owners or our team will answer.

Original document

Emerson Tupelo pH Sensor (396) manual PDF

12 pages·647 KB·Photo Scanner·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…