20002158, Rev DI March 2014


Safety and approval information
This Micro Motion product complies with all applicable European directives when properly installed in accordance with the instructions in this manual. Refer to the EC declaration of conformity for directives that apply to this product. The EC declaration of conformity, with all applicable European directives, and the complete ATEX Installation Drawings and Instructions are available on the internet at www.micromotion.com or through your local Micro Motion support center.
Information affixed to equipment that complies with the Pressure Equipment Directive can be found on the internet at www.micromotion.com/documentation.
For hazardous installations in Europe, refer to standard EN 60079-14 if national standards do not apply.
Full product specifications can be found in the product data sheet. Troubleshooting information can be found in the transmitter configuration manual. Product data sheets and manuals are available from the Micro Motion web site at www.micromotion.com/documentation.
Micro Motion procedures must be followed when returning equipment. These procedures ensure legal compliance with government transportation agencies and help provide a safe working environment for Micro Motion employees. Failure to follow Micro Motion procedures will result in your equipment being refused delivery.
Information on return procedures and forms is available on our web support system at www.micromotion.com, or by phoning the Micro Motion Customer Service department.
Micro Motion customer service
Email:
Telephone:

Other information · Return policy


□ Make sure that the hazardous area specified on the approval tag is suitable for the environment in which the meter will be installed. □ Verify that the local ambient and process temperatures are within the limits of the meter. □ If your sensor has an integral transmitter, no wiring is required between the sensor and transmitter. Follow the wiring instructions in the transmitter installation manual for signal and power wiring. □ If your transmitter has remote-mounted electronics, follow the instructions in this manual for wiring between the sensor and the transmitter, and then follow the instructions in the transmitter installation manual for power and signal wiring. □ For the wiring between the sensor and the transmitter, consider maximum cable lengths. The maximum distance between the sensor and transmitter depends on the cable type. For all types of wiring, Micro Motion recommends using Micro Motion cable.
Planning

□ For optimal performance, install the sensor in the preferred orientation. The sensor will work in any orientation as long as the flow tubes remain full of process fluid.

□ Install the meter so that the flow direction arrow on the sensor case matches the actual forward flow of the process. (Flow direction is also software-selectable.)
The following information can help you get the most from your sensor.
Planning

Planning

Ambient and process temperature limits for ELITE CMF***M/L/H/P (excludes cryogenic modifications) and CMFS007-015 meters Figure 1-1:
D
A. Ambient temperature °F (°C) B. Process temperature in °F (°C) C. All available electronic options D. Extended or remote mount electronics only
D
D
A. Ambient temperature °F (°C) B. Process temperature in °F (°C) C. All available electronic options D. Extended or remote mount electronics only

D
A. Ambient temperature °F (°C) B. Process temperature in °F (°C) C. All available electronic options D. Extended or remote mount electronics only
D
A. Ambient temperature °F (°C) B. Process temperature in °F (°C) C. All available electronic options D. Extended or remote mount electronics only

D
A. Ambient temperature °F (°C) B. Process temperature in °F (°C) C. All available electronic options D. Extended or remote mount electronics only
CMFS sensors are certified EHEDG TYPE EL CLASS I for hygienic applications when installed vertically with the process fitting and gasket combinations listed in the Position Paper of the EHEDG Test Methods Subgroup (available at http://www.ehedg.org). Other process connections/gasket combinations may be used provided they have been evaluated and successfully tested for in-place cleanability according to the latest edition of EHEDG Document 2. Refer to the ELITE Product Data Sheet for further information about fitting options.
For optimal cleanability and drainability:

A. Process pipeline B. Direction of normal process flow C. Direction of drainage
Planning

Heavy meters (those over 50 lb. [22 kg]), and even lighter meters that must be installed in elevated or difficult-to-reach places, usually require additional consideration when transporting or lifting them into their installation location.
- Fixed hoist boom trucks or cranes
- Continuous web belt slings
- Eye to eye web belt slings
- Two leg wire rope slings


A. Typical center of gravity
Use your common practices to minimize torque and bending load on process connections.

Do not lift the sensor by the electronics or purge connections. Lifting the sensor by the electronics or purge connections can damage the device.
The electronics of high-temperature sensors are attached to the end of a 32" (812 mm) pre-installed flexible conduit. The electronics must be separately mounted on a wall or instrument pole.


A. Sensor B. Electronics C. Mounting bracket D. Flexible conduit (minimum bend radius 6" [152 mm])
With some large meter sizes, the meter may ship with the electronics attached to the sensor case. The meter cannot be operated in this configuration. Detach the electronics bracket from the sensor case and then proceed to mount the electronics to a wall or instrument pole as described below.


A. Detach electronics from sensor case and mount to a wall or instrument pole









A. Wall or flat surface B. Electronics (enhanced core processor shown) C. Flexible conduit


A. Instrument pole B. Electronics (enhanced core processor shown) C. Flexible conduit
The CMF010 sensor has an optional mounting configuration for use with small or flexible pipeline. If the pipeline adequately supports the sensor, this procedure can be skipped.


A. Two user-supplied 5/16" (M8) bolts B. Junction box or core processor (junction box shown) C. Mounting surface
The CMFS007, CMFS010 and CMFS015 sensors have an optional mounting bracket for use with small or flexible pipeline. If the pipeline adequately supports the sensor, this procedure can be skipped.


A. Mounting bracket B. Mounting holes C. Supplied U-bolts
2.6 Mount a CMFS025, CMFS040 or CMFS050 sensor in a wall mount bracket
The CMFS025, CMFS040, and CMFS050 sensors have an optional wall mounting bracket.



A wafer-style connection lets you clamp the sensor into the pipeline. A wafer installation kit is shipped with a wafer-style sensor.


A. Flange nut B. Gasket (user-supplied) C. Alignment ring D. Flange bolt E. Sensor wafer
B A





A. Direction to rotate the alignment ring B. Direction the flange bolts are pushed C. Flange bolt

If you ordered a sensor with extended electronics, you will need to install the extender onto the sensor case.
Keep the extender and feedthrough clean and dry. Moisture or debris in the extender or feedthrough can damage electronics and result in measurement error or flowmeter failure.


D
F
A. Transmitter or core processor B. Extender C. O-ring D. Feedthrough E. Clamping ring F. Clamping screw G. Plastic plug H. Plastic cap

The wiring procedure you follow depends on which electronics option you have.
Electronics option Wiring procedure
Integral transmitter No wiring is required between sensor and transmitter. See the transmitter installation manual for wiring the power and signal cable to the transmitter.
™ No transmitter to wire. See the MVD Direct Connect manual for wiring the power and signal cable between the sensor and the direct host.
Core processor Connect a 4-wire cable between the sensor and transmitter. Refer to the 4-wire connection instructions in this manual.
Junction box Connect a 9-wire cable between the sensor and transmitter. Refer to the 9-wire connection instructions in this manual, as well as the Micro Motion 9-Wire Flowmeter Cable Preparation and Installation Guide.
Make sure the hazardous area specified on the sensor approval tag is suitable for the environment in which the sensor will be installed. Failure to comply with the requirements for intrinsic safety in a hazardous area could result in an explosion.
Fully close and tighten all housing covers and conduit openings. Improperly sealed housings can expose electronics to moisture, which can cause measurement error or flowmeter failure. Inspect and grease all gaskets and O-rings.
Wiring · ™Direct Connect

3.2.1 Prepare the 4-wire cable
Wiring

Remove the core processor
Pass the wires through the gland nut and clamping insert.
Lay cable in conduit
Pass the wires through the gland. Terminate the drain

gland.
Done (do not perform the shielding procedure)
Wrap the drain wires twice around the shield and cut off
the excess drain wires.

Drain wires wrapped around shield
Go to the shielding

Gland supplier Micro Motion
type Braided (armored cable)
Trim 7 mm from the shielded
Apply the Heat Shrink 1. Slide the shielded heat shrink over the drain wires. Ensure that the wires are completely covered. 2. Apply heat (250 °F or 120 °C) to shrink the tubing. Do not burn the cable. 3. Position the clamping insert so the interior end is flush with the braid of the heat shrink.

Terminate the shield and drain wires in the
Assemble the Gland 1. Fold the shield or braid back over the clamping insert and 1/8 inch (3 mm) past the O-ring. 2. Install the gland body into the conduit opening on the core processor housing. 3. Insert the wires through gland body and tighten the gland nut onto the gland body.
Assemble the gland according to vendor

Micro Motion offers two types of 4-wire cable: shielded and armored. Both types contain shield drain wires.
The 4-wire cable supplied by Micro Motion consists of one pair of red and black 18 AWG (0.75 mm2) wires for the VDC connection, and one pair of white and green 22 AWG (0.35 mm2) wires for the RS-485 connection.
User-supplied 4-wire cable must meet the following requirements:

Wire gauge Maximum cable length
VDC 22 AWG (0.35 mm2) 300 ft (90 m)
VDC 20 AWG (0.5 mm2) 500 ft (150 m)
VDC 18 AWG (0.8 mm2) 1000 ft (300 m)
RS-485 22 AWG (0.35 mm2) or larger 1000 ft (300 m)
After the 4-wire cable has been prepared and shielded (if required), connect the individual wires of the 4-wire cable to the terminals on the core processor.
Wiring

From Step 1 or 2
Enhanced core processor Standard core processor
Connect the wires to the core processor terminals: Red wire > Terminal 1 (Power supply +) Black wire > Terminal 2 (Power supply –) White wire > Terminal 3 (RS-485/A) Green wire > Terminal 4 (RS-485/B)
Connect the wires to the core processor terminals: Red wire > Terminal 1 (Power supply +) Black wire > Terminal 2 (Power supply –) White wire > Terminal 3 (RS-485/A) Green wire > Terminal 4 (RS-485/B)
Reinstall and tighten the core processor cover
Connect the wires to the transmitter
terminals (see the transmitter manual)
Wiring · Core processor · type

Wiring

The meter must be grounded according to the standards that are applicable at the site. The customer is responsible for knowing and complying with all applicable standards.
Micro Motion suggests the following guides for grounding practices:
If no external standards are applicable, follow these guidelines to ground the sensor:
Ground the flowmeter to earth, or follow ground network requirements for the facility. Improper grounding can cause measurement error.
Check the joints in the pipeline.
- If the joints in the pipeline are ground-bonded, the sensor is automatically grounded and no further action is necessary (unless required by local code).
- If the joints in the pipeline are not grounded, connect a ground wire to the grounding screw located on the sensor electronics.
Grounding

If the sensor has purge fittings, they should remain sealed at all times. After a purge plug has been removed, the sensor case should be purged with argon or nitrogen and resealed.
Purging the case protects internal components. The sensor is purged of all oxygen and sealed at the factory. If the purge plugs are never removed, it is not necessary to purge or re-seal the sensor. For more information, contact Micro Motion Customer Service.
If a purge plug is removed from the sensor case, it will be necessary to repurge the case.
Take all necessary precautions when removing purge plugs. Removing a purge plug compromises the secondary containment of the sensor and could expose the user to process fluid.
Improper pressurization of the sensor case could result in personal injury. Removing a purge plug will require the sensor case to be repurged with a dry inert gas. Follow all instructions provided in the case purging procedure.
Make sure the following are available before beginning the purge procedure:
Supplementary information

The purge time is the amount of time required for full exchange of atmosphere to inert gas. The larger the line size, the greater amount of time is required to purge the case. If purge lines are being used, increase the purge time to fill the additional volume of the purge line.

Sensor model Purge rate, in ft3/hr (l/h) Time, in minutes
Supplementary information

Sensor model Purge rate, in ft3/hr (l/h) Time, in minutes
CMFS150 20 (566) 6
Rupture disks are meant to vent process fluid from the sensor case in the event of a flow tube rupture. Some users connect a pipeline to the rupture disk to help contain escaping process fluid. For more information about rupture disks, contact Micro Motion Customer Service.
If the sensor has rupture disks, they are installed in the sensor purge fitting openings. The rupture disks should remain installed at all times. If you remove a rupture disk from the sensor case, it will be necessary to re-purge the case.
Stay clear of the rupture disk pressure relief area. High-pressure fluid escaping from the sensor can cause severe injury or death.
Supplementary information

Micro Motion Inc. USA Worldwide Headquarters 7070 Winchester Circle Boulder, Colorado 80301 T +1 303-527-5200 T +1 800-522-6277 F +1 303-530-8459 www.micromotion.com
Micro Motion Europe Emerson Process Management Neonstraat 1 6718 WX Ede The Netherlands T +31 (0) 318 495 555 F +31 (0) 318 495 556 www.micromotion.nl
Micro Motion Asia Emerson Process Management 1 Pandan Crescent Singapore 128461 Republic of Singapore T +65 6777-8211 F +65 6770-8003
Micro Motion United Kingdom Emerson Process Management Limited Horsfield Way Bredbury Industrial Estate Stockport SK6 2SU U.K. T +44 0870 240 1978 F +44 0800 966 181
Micro Motion Japan Emerson Process Management 1-2-5, Higashi Shinagawa Shinagawa-ku Tokyo 140-0002 Japan T +81 3 5769-6803 F +81 3 5769-6844
©2014 Micro Motion, Inc. All rights reserved.