Page 1 of the Emerson Thermostat (20002158) user manual — Micro Motion ® ELITE ® Coriolis Flow and Density Sensors
Page 1 · click to enlarge
Page 1

20002158, Rev DI March 2014

Micro Motion ® ELITE ® Coriolis Flow and Density Sensors

Emerson Thermostat (20002158) — Micro Motion ® ELITE ® Coriolis Flow and Density Sensors diagram (page 1)
Fig. 1 · Micro Motion ® ELITE ® Coriolis Flow and Density Sen…
Page 2 of the Emerson Thermostat (20002158) user manual — Safety and approval information
Page 2 · click to enlarge
Page 2

Safety and approval information

This Micro Motion product complies with all applicable European directives when properly installed in accordance with the instructions in this manual. Refer to the EC declaration of conformity for directives that apply to this product. The EC declaration of conformity, with all applicable European directives, and the complete ATEX Installation Drawings and Instructions are available on the internet at www.micromotion.com or through your local Micro Motion support center.

Information affixed to equipment that complies with the Pressure Equipment Directive can be found on the internet at www.micromotion.com/documentation.

For hazardous installations in Europe, refer to standard EN 60079-14 if national standards do not apply.

Full product specifications can be found in the product data sheet. Troubleshooting information can be found in the transmitter configuration manual. Product data sheets and manuals are available from the Micro Motion web site at www.micromotion.com/documentation.

Micro Motion procedures must be followed when returning equipment. These procedures ensure legal compliance with government transportation agencies and help provide a safe working environment for Micro Motion employees. Failure to follow Micro Motion procedures will result in your equipment being refused delivery.

Information on return procedures and forms is available on our web support system at www.micromotion.com, or by phoning the Micro Motion Customer Service department.

Micro Motion customer service

Email:

  • Worldwide: flow.support@<a href="http://emerson.com" rel="nofollow noopener" target="_blank">emerson.com</a>
  • Asia-Pacific: APflow.support@<a href="http://emerson.com" rel="nofollow noopener" target="_blank">emerson.com</a>

Telephone:

North and South AmericaEurope and Middle EastAsia Pacific United States800-522-6277U.K.0870 240 1978Australia800 158 727 Canada+1 303-527-5200The Netherlands+31 (0) 318 495 555New Zealand099 128 804 Mexico+41 (0) 41 7686 111France0800 917 901India800 440 1468 Argentina+54 11 4837 7000Germany0800 182 5347Pakistan888 550 2682 Brazil+55 15 3413 8000Italy8008 77334China+86 21 2892 9000 Venezuela+58 26 1731 3446Central & Eastern+41 (0) 41 7686 111Japan+81 3 5769 6803 Russia/CIS+7 495 981 9811South Korea+82 2 3438 4600 Egypt0800 000 0015Singapore+65 6 777 8211 Oman800 70101Thailand001 800 441 6426 Qatar431 0044Malaysia800 814 008 Kuwait663 299 01 South Africa800 991 390 Saudia Arabia800 844 9564 UAE800 0444 0684
Emerson Thermostat (20002158) — Safety and approval information diagram (page 2)
Fig. 2 · Safety and approval information

Other information · Return policy

Page 5 of the Emerson Thermostat (20002158) user manual — Maximum lengths for user-supplied 4-wire cable Table 1-2:
Page 5 · click to enlarge
Page 5

1 Planning

Topics covered in this chapter:

  • Installation checklist
  • Best practices
  • Temperature limits
  • Recommendations for hygienic and self-draining applications

1.1 Installation checklist

□ Make sure that the hazardous area specified on the approval tag is suitable for the environment in which the meter will be installed. □ Verify that the local ambient and process temperatures are within the limits of the meter. □ If your sensor has an integral transmitter, no wiring is required between the sensor and transmitter. Follow the wiring instructions in the transmitter installation manual for signal and power wiring. □ If your transmitter has remote-mounted electronics, follow the instructions in this manual for wiring between the sensor and the transmitter, and then follow the instructions in the transmitter installation manual for power and signal wiring. □ For the wiring between the sensor and the transmitter, consider maximum cable lengths. The maximum distance between the sensor and transmitter depends on the cable type. For all types of wiring, Micro Motion recommends using Micro Motion cable.

Maximum lengths for Micro Motion cable Table 1-1:

Cable typeTo transmitterMaximum length Micro Motion 9-wire9739 MVD transmitter1000 ft (300 m) All other MVD transmitters60 ft (20 m) Micro Motion 4-wireAll 4-wire MVD transmitters1000 ft (300 m)

Maximum lengths for user-supplied 4-wire cable Table 1-2:

Wire functionWire sizeMaximum length Power (VDC)22 AWG (0,35 mm2)300 ft (90 m) 20 AWG (0,5 mm2)500 ft (150 m) 18 AWG (0,8 mm2)1000 ft (300 m) Signal (RS-485)22 AWG (0,35 mm2) or larger1000 ft (300 m)

Planning

Page 6 of the Emerson Thermostat (20002158) user manual — 1.2 Best practices
Page 6 · click to enlarge
Page 6

□ For optimal performance, install the sensor in the preferred orientation. The sensor will work in any orientation as long as the flow tubes remain full of process fluid.

Preferred sensor orientation Table 1-3:

ProcessPreferred orientationAlternate orientations Liquids Gases Slurries
Emerson Thermostat (20002158) — Preferred sensor orientation Table 1-3: diagram (page 6)
Fig. 3 · Preferred sensor orientation Table 1-3:

□ Install the meter so that the flow direction arrow on the sensor case matches the actual forward flow of the process. (Flow direction is also software-selectable.)

1.2 Best practices

The following information can help you get the most from your sensor.

Planning

Page 7 of the Emerson Thermostat (20002158) user manual — 1.3 Temperature limits
Page 7 · click to enlarge
Page 7
  • There are no pipe run requirements for Micro Motion sensors. Straight runs of pipe upstream or downstream are unnecessary.
  • If the sensor is installed in a vertical pipeline, liquids and slurries should flow upward through the sensor. Gases may flow upward or downward.
  • Keep the sensor tubes full of process fluid.
  • For halting flow through the sensor with a single valve, install the valve downstream from the sensor.
  • Minimize bending and torsional stress on the meter. Do not use the meter to align misaligned piping.
  • The sensor does not require external supports. The flanges will support the sensor in any orientation. (Some sensor models installed in very small, flexible pipeline have optional installation instructions that allow for external supports.)

1.3 Temperature limits

  • Sensors can be used in the process and ambient temperature ranges shown in the temperature limit graphs. For the purposes of selecting electronics options, temperature limit graphs should be used only as a general guide. If your process conditions are close to the gray areas, it may be inappropriate to use electronics options other than a junction box. Consult consult with your Micro Motion representative.
  • In all cases, the electronics cannot be operated where the ambient temperature is below –40 °F (–40 °C) or above +140 °F (+60 °C). If a sensor is to be used where the ambient temperature is outside of the range permissible for the electronics, the electronics must be remotely located where the ambient temperature is within the permissible range, as indicated by the shaded area of the temperature limit graphs.
  • Temperature limits may be further restricted by hazardous area approvals. Refer to the hazardous area approvals documentation shipped with the sensor or available from the Micro Motion web site ( <a href="http://www.micromotion.com" rel="nofollow noopener" target="_blank">www.micromotion.com</a> ).
  • The extended-mount electronics option allows the sensor case to be insulated without covering the transmitter, core processor, or junction box, but does not affect temperature ratings.

Planning

Page 8 of the Emerson Thermostat (20002158) user manual — Ambient and process temperature limits for ELITE CMFS025–CMFS150 meters Figure 1-2:
Page 8 · click to enlarge
Page 8

Ambient and process temperature limits for ELITE CMF***M/L/H/P (excludes cryogenic modifications) and CMFS007-015 meters Figure 1-1:

D

D

A. Ambient temperature °F (°C) B. Process temperature in °F (°C) C. All available electronic options D. Extended or remote mount electronics only

Ambient and process temperature limits for ELITE CMFS025–CMFS150 meters Figure 1-2:

D

D

A. Ambient temperature °F (°C) B. Process temperature in °F (°C) C. All available electronic options D. Extended or remote mount electronics only

Show 15 more labels / short text on this page
Planning · 140 (60) · 140 (60) · 113 (45) · –40 (–40) · –148 (–100) · –400 (–240) · 400 (204) · 140 (60) · 140 (60) · 113 (45) · –40 (–40) · –148 (–100) · –58 (–50) · 400 (204)
Page 9 of the Emerson Thermostat (20002158) user manual — Ambient and process temperature limits for high temperature ELITE meters Figure 1-4:
Page 9 · click to enlarge
Page 9

Ambient and process temperature limits for cryogenic ELITE meters Figure 1-3:

D

A. Ambient temperature °F (°C) B. Process temperature in °F (°C) C. All available electronic options D. Extended or remote mount electronics only

Ambient and process temperature limits for high temperature ELITE meters Figure 1-4:

D

A. Ambient temperature °F (°C) B. Process temperature in °F (°C) C. All available electronic options D. Extended or remote mount electronics only

Show 11 more labels / short text on this page
Planning · 140 (60) · –40 (–40) · –148 (–100) · –400 (–240) · 176 (80) · 140 (60) · –40 (–40) · –148 (–100) · –50 (–58) · 662 (350)
Page 10 of the Emerson Thermostat (20002158) user manual — 1.4 Recommendations for hygienic and selfdraining applications
Page 10 · click to enlarge
Page 10

Ambient and process temperature limits for Super Duplex ELITE meters Figure 1-5:

D

A. Ambient temperature °F (°C) B. Process temperature in °F (°C) C. All available electronic options D. Extended or remote mount electronics only

1.4 Recommendations for hygienic and selfdraining applications

CMFS sensors are certified EHEDG TYPE EL CLASS I for hygienic applications when installed vertically with the process fitting and gasket combinations listed in the Position Paper of the EHEDG Test Methods Subgroup (available at http://www.ehedg.org). Other process connections/gasket combinations may be used provided they have been evaluated and successfully tested for in-place cleanability according to the latest edition of EHEDG Document 2. Refer to the ELITE Product Data Sheet for further information about fitting options.

For optimal cleanability and drainability:

  • If possible, install the sensor in a vertical pipeline with the process fluid flowing upward through the sensor.
  • If the sensor must be installed in a horizontal pipeline, drainage is accomplished by air purge evacuation of the pipeline circuit.
  • For clean-in-place (CIP) applications, Micro Motion recommends using the generally-accepted flow velocity of at least 1.5 m/s for cleaning the sensor.
  • The gap between the electronics housing and sensor body should be inspected periodically. Manually clean this gap when necessary.
Show 6 more labels / short text on this page
Planning · 140 (60) · –40 (–40) · –148 (–100) · –40 (–40) · 400 (204)
Page 11 of the Emerson Thermostat (20002158) user manual — Installation for self-draining applications Figure 1-6:
Page 11 · click to enlarge
Page 11

Installation for self-draining applications Figure 1-6:

A. Process pipeline B. Direction of normal process flow C. Direction of drainage

Planning

Page 12 of the Emerson Thermostat (20002158) user manual — 2.1 Recommendations for lifting heavy meters
Page 12 · click to enlarge
Page 12

Topics covered in this chapter:

  • Recommendations for lifting heavy meters
  • Mount the sensor
  • Mount electronics of high-temperature sensors
  • Mount a CMF010 sensor to a wall or pole
  • Mount a CMFS007, CMFS010 or CMFS015 sensor in a bracket
  • Mount a CMFS025, CMFS040 or CMFS050 sensor in a wall mount bracket
  • Secure wafer-style process connections
  • Attach extended electronics

2.1 Recommendations for lifting heavy meters

Heavy meters (those over 50 lb. [22 kg]), and even lighter meters that must be installed in elevated or difficult-to-reach places, usually require additional consideration when transporting or lifting them into their installation location.

  • Safe handling during transportation and installation is the responsibility of the installer – know and follow all safety practices and regulations for your facility and for any lifting/rigging equipment being used
  • A professional rigging crew with proper equipment should be used
  • Typical equipment for handling heavy meters includes the following:

- Fixed hoist boom trucks or cranes

- Continuous web belt slings

- Eye to eye web belt slings

- Two leg wire rope slings

  • Lift a meter by its case or on the case side of its process fittings.
  • Do not lift a meter by its electronics (junction box, transmitter, or any electrical fittings) or by its purge fittings
  • It may be useful to identify the meter center of gravity
  • Protect the sealed surfaces on the process fittings with factory-installed flange protectors or comparable field-installed protection
Page 13 of the Emerson Thermostat (20002158) user manual — 2.2 Mount the sensor
Page 13 · click to enlarge
Page 13

Acceptable lifting points Figure 2-1:

Emerson Thermostat (20002158) — Acceptable lifting points Figure 2-1: diagram (page 13)
Fig. 4 · Acceptable lifting points Figure 2-1:

Center of gravity for large meters Figure 2-2:

A. Typical center of gravity

2.2 Mount the sensor

Use your common practices to minimize torque and bending load on process connections.

Page 14 of the Emerson Thermostat (20002158) user manual — 2.3 Mount electronics of high-temperature sensors
Page 14 · click to enlarge
Page 14

CAUTION!

Do not lift the sensor by the electronics or purge connections. Lifting the sensor by the electronics or purge connections can damage the device.

Mounting the sensor Figure 2-3:

  • Notes
  • Do not use the sensor to support the piping.
  • The sensor does not require external supports. The flanges will support the sensor in any orientation. (Some sensor models installed in very small, flexible pipeline have optional installation instructions that allow for external supports.)

2.3 Mount electronics of high-temperature sensors

The electronics of high-temperature sensors are attached to the end of a 32" (812 mm) pre-installed flexible conduit. The electronics must be separately mounted on a wall or instrument pole.

Page 15 of the Emerson Thermostat (20002158) user manual — D
Page 15 · click to enlarge
Page 15

Components of a high-temperature sensor Figure 2-4:

Emerson Thermostat (20002158) — Components of a high-temperature sensor Figure 2-4: diagram (page 15)
Fig. 5 · Components of a high-temperature sensor Figure 2-4:

D

A. Sensor B. Electronics C. Mounting bracket D. Flexible conduit (minimum bend radius 6" [152 mm])

With some large meter sizes, the meter may ship with the electronics attached to the sensor case. The meter cannot be operated in this configuration. Detach the electronics bracket from the sensor case and then proceed to mount the electronics to a wall or instrument pole as described below.

Page 16 of the Emerson Thermostat (20002158) user manual — Procedure
Page 16 · click to enlarge
Page 16

Removing electronics from the sensor case Figure 2-5:

Emerson Thermostat (20002158) — Removing electronics from the sensor case Figure 2-5: diagram (page 16)
Fig. 6 · Removing electronics from the sensor case Figure 2-5…

A. Detach electronics from sensor case and mount to a wall or instrument pole

Procedure

  • For wall mounting, use four 5/16" or four M8 bolts to secure the mounting bracket.
Page 17 of the Emerson Thermostat (20002158) user manual — Wall-mount components Figure 2-6:
Page 17 · click to enlarge
Page 17

Wall-mount components Figure 2-6:

Emerson Thermostat (20002158) — Wall-mount components Figure 2-6: diagram (page 17)
Fig. 7 · Wall-mount components Figure 2-6:
Emerson Thermostat (20002158) — Wall-mount components Figure 2-6: diagram (page 17)
Fig. 8 · Wall-mount components Figure 2-6:
Emerson Thermostat (20002158) — Wall-mount components Figure 2-6: diagram (page 17)
Fig. 9 · Wall-mount components Figure 2-6:
Emerson Thermostat (20002158) — Wall-mount components Figure 2-6: diagram (page 17)
Fig. 10 · Wall-mount components Figure 2-6:
Emerson Thermostat (20002158) — Wall-mount components Figure 2-6: diagram (page 17)
Fig. 11 · Wall-mount components Figure 2-6:
Emerson Thermostat (20002158) — Wall-mount components Figure 2-6: diagram (page 17)
Fig. 12 · Wall-mount components Figure 2-6:
Emerson Thermostat (20002158) — Wall-mount components Figure 2-6: diagram (page 17)
Fig. 13 · Wall-mount components Figure 2-6:
Emerson Thermostat (20002158) — Wall-mount components Figure 2-6: diagram (page 17)
Fig. 14 · Wall-mount components Figure 2-6:

A. Wall or flat surface B. Electronics (enhanced core processor shown) C. Flexible conduit

  • For mounting to an instrument pole, use a 2-inch U-bolt pipe kit to secure the mounting bracket.
Page 18 of the Emerson Thermostat (20002158) user manual — 2.4 Mount a CMF010 sensor to a wall or pole
Page 18 · click to enlarge
Page 18

Pole-mount components Figure 2-7:

Emerson Thermostat (20002158) — Pole-mount components Figure 2-7: diagram (page 18)
Fig. 15 · Pole-mount components Figure 2-7:

A. Instrument pole B. Electronics (enhanced core processor shown) C. Flexible conduit

2.4 Mount a CMF010 sensor to a wall or pole

The CMF010 sensor has an optional mounting configuration for use with small or flexible pipeline. If the pipeline adequately supports the sensor, this procedure can be skipped.

1. Locate the optional mounting holes. For sensors with a junction box, the junction box must be rotated to the side to expose the mounting holes.
Page 19 of the Emerson Thermostat (20002158) user manual — 2.5 Mount a CMFS007, CMFS010 or CMFS015 sensor in a bracket
Page 19 · click to enlarge
Page 19

Optional mounting for CMF010 sensors Figure 2-8:

Emerson Thermostat (20002158) — Optional mounting for CMF010 sensors Figure 2-8: diagram (page 19)
Fig. 16 · Optional mounting for CMF010 sensors Figure 2-8:

A. Two user-supplied 5/16" (M8) bolts B. Junction box or core processor (junction box shown) C. Mounting surface

2. If necessary, install rigid standoffs between the sensor and the mounting surface.
3. Using two user-supplied 5/16" (M8) bolts (minimum length 2 1/4" [58 mm]), secure the sensor case to the mounting surface.

2.5 Mount a CMFS007, CMFS010 or CMFS015 sensor in a bracket

The CMFS007, CMFS010 and CMFS015 sensors have an optional mounting bracket for use with small or flexible pipeline. If the pipeline adequately supports the sensor, this procedure can be skipped.

1. Secure the mounting bracket to a wall or other flat surface with four user-supplied 5/16" (8 mm) bolts.
2. Place the sensor into the bracket.
3. Secure the sensor in the bracket with the supplied 5/16" (M8) U-bolts.
Page 20 of the Emerson Thermostat (20002158) user manual — Assembled wall mounting bracket for CMFS025, CMFS040, and CMFS050 Figure 2-10:
Page 20 · click to enlarge
Page 20

Mounting bracket for CMFS007, CMFS010, and CMFS015 Figure 2-9:

Emerson Thermostat (20002158) — Mounting bracket for CMFS007, CMFS010, and CMFS015 Figure 2-9: diagram (page 20)
Fig. 17 · Mounting bracket for CMFS007, CMFS010, and CMFS015 F…

A. Mounting bracket B. Mounting holes C. Supplied U-bolts

2.6 Mount a CMFS025, CMFS040 or CMFS050 sensor in a wall mount bracket

The CMFS025, CMFS040, and CMFS050 sensors have an optional wall mounting bracket.

1. Assemble the bracket.

Assembled wall mounting bracket for CMFS025, CMFS040, and CMFS050 Figure 2-10:

Emerson Thermostat (20002158) — Assembled wall mounting bracket for CMFS025, CMFS040, and CMFS050 Figure 2-10: diagram (page 20)
Fig. 18 · Assembled wall mounting bracket for CMFS025, CMFS040…
2. Attach the bracket to the wall using fasteners appropriate for the mounting surface.
Page 21 of the Emerson Thermostat (20002158) user manual — 2.7 Secure wafer-style process connections
Page 21 · click to enlarge
Page 21
3. Place the sensor into the bracket.
4. Secure the sensor in the bracket with the supplied fasteners.

CMFS025, CMFS040, or CMFS050 wall mounted using bracket Figure 2-11:

Emerson Thermostat (20002158) — CMFS025, CMFS040, or CMFS050 wall mounted using bracket Figure 2-11: diagram (page 21)
Fig. 19 · CMFS025, CMFS040, or CMFS050 wall mounted using brac…

2.7 Secure wafer-style process connections

A wafer-style connection lets you clamp the sensor into the pipeline. A wafer installation kit is shipped with a wafer-style sensor.

1. Make sure that the bolts provided are rated for your process connection.
2. Slip the sensor alignment rings over each end of the sensor wafer, then insert the sensor between the process connections in the pipeline.
Page 22 of the Emerson Thermostat (20002158) user manual — Alignment ring usage Figure 2-13:
Page 22 · click to enlarge
Page 22

Wafer-style connection components Figure 2-12:

Emerson Thermostat (20002158) — Wafer-style connection components Figure 2-12: diagram (page 22)
Fig. 20 · Wafer-style connection components Figure 2-12:

A. Flange nut B. Gasket (user-supplied) C. Alignment ring D. Flange bolt E. Sensor wafer

3. Insert the flange bolts through both process connections, and thread the flange nuts onto the bolts.
4. With your fingers, tighten the flange nuts.
5. Rotate the sensor alignment rings in the direction that pushes the bolts outward.

Alignment ring usage Figure 2-13:

B A

Emerson Thermostat (20002158) — Alignment ring usage Figure 2-13: diagram (page 22)
Fig. 21 · Alignment ring usage Figure 2-13:
Emerson Thermostat (20002158) — Alignment ring usage Figure 2-13: diagram (page 22)
Fig. 22 · Alignment ring usage Figure 2-13:
Emerson Thermostat (20002158) — Alignment ring usage Figure 2-13: diagram (page 22)
Fig. 23 · Alignment ring usage Figure 2-13:
Emerson Thermostat (20002158) — Alignment ring usage Figure 2-13: diagram (page 22)
Fig. 24 · Alignment ring usage Figure 2-13:
Emerson Thermostat (20002158) — Alignment ring usage Figure 2-13: diagram (page 22)
Fig. 25 · Alignment ring usage Figure 2-13:

A. Direction to rotate the alignment ring B. Direction the flange bolts are pushed C. Flange bolt

Page 23 of the Emerson Thermostat (20002158) user manual — Procedure
Page 23 · click to enlarge
Page 23
6. With a wrench, tighten the nuts in an alternating order.

2.8 Attach extended electronics

If you ordered a sensor with extended electronics, you will need to install the extender onto the sensor case.

CAUTION!

Keep the extender and feedthrough clean and dry. Moisture or debris in the extender or feedthrough can damage electronics and result in measurement error or flowmeter failure.

Procedure

1. Remove and recycle the plastic cap from the feedthrough on the sensor.
Page 24 of the Emerson Thermostat (20002158) user manual — G
Page 24 · click to enlarge
Page 24

Feedthrough and extender components Figure 2-14:

Emerson Thermostat (20002158) — Feedthrough and extender components Figure 2-14: diagram (page 24)
Fig. 26 · Feedthrough and extender components Figure 2-14:

G

D

F

A. Transmitter or core processor B. Extender C. O-ring D. Feedthrough E. Clamping ring F. Clamping screw G. Plastic plug H. Plastic cap

2. Loosen the clamping screw and remove the clamping ring. Leave the O-ring in place on the feedthrough.
3. Remove and recycle the plastic plug from the extender.
4. Fit the extender onto the feedthrough by carefully aligning the notches on the bottom of the extender with the notches on the feedthrough.
5. Close the clamping ring and tighten the clamping screw to 13–18 in-lbs. (1,5–2 N-m).
Page 25 of the Emerson Thermostat (20002158) user manual — CAUTION!
Page 25 · click to enlarge
Page 25

3 Wiring

Topics covered in this chapter:

  • Options for wiring
  • Connect 4-wire cable
  • Connect 9-wire cable

3.1 Options for wiring

The wiring procedure you follow depends on which electronics option you have.

Wiring procedures by electronics option Table 3-1:

Electronics option Wiring procedure

Integral transmitter No wiring is required between sensor and transmitter. See the transmitter installation manual for wiring the power and signal cable to the transmitter.

MVD

™ No transmitter to wire. See the MVD Direct Connect manual for wiring the power and signal cable between the sensor and the direct host.

Core processor Connect a 4-wire cable between the sensor and transmitter. Refer to the 4-wire connection instructions in this manual.

Junction box Connect a 9-wire cable between the sensor and transmitter. Refer to the 9-wire connection instructions in this manual, as well as the Micro Motion 9-Wire Flowmeter Cable Preparation and Installation Guide.

CAUTION!

Make sure the hazardous area specified on the sensor approval tag is suitable for the environment in which the sensor will be installed. Failure to comply with the requirements for intrinsic safety in a hazardous area could result in an explosion.

CAUTION!

Fully close and tighten all housing covers and conduit openings. Improperly sealed housings can expose electronics to moisture, which can cause measurement error or flowmeter failure. Inspect and grease all gaskets and O-rings.

Wiring · ™Direct Connect

Page 26 of the Emerson Thermostat (20002158) user manual — 3.2 Connect 4-wire cable
Page 26 · click to enlarge
Page 26

3.2 Connect 4-wire cable

3.2.1 Prepare the 4-wire cable

Wiring

Page 27 of the Emerson Thermostat (20002158) user manual — NPT M20
Page 27 · click to enlarge
Page 27

4-wire cable preparation Figure 3-1:

Remove the core processor

Pass the wires through the gland nut and clamping insert.

Lay cable in conduit

Pass the wires through the gland. Terminate the drain

Emerson Thermostat (20002158) — 4-wire cable preparation Figure 3-1: diagram (page 27)
Fig. 27 · 4-wire cable preparation Figure 3-1:

gland.

Done (do not perform the shielding procedure)

NPT M20

1. Strip 4-1/2 inch (115 mm) of cable jacket. 2. Remove the clear wrap and filler material. 3. Strip all but 3/4 inch (19 mm) of shielding.
1. Strip 4-1/4 inch (108 mm) of cable jacket. 2. Remove the clear wrap and filler material. 3. Strip all but 1/2 inch (12 mm) of shielding.

Wrap the drain wires twice around the shield and cut off

the excess drain wires.

Emerson Thermostat (20002158) — NPT M20 diagram (page 27)
Fig. 28

Drain wires wrapped around shield

Go to the shielding

Show 17 more labels / short text on this page
Wiring · cover · Cable layout · Cable glands · Metal conduit · Gland supplier · Micro Motion · Run conduit to · User-supplied · sensor · cable gland · cable gland · Gland nut · Clamping insert · wires inside the · Gland type · procedure
Page 28 of the Emerson Thermostat (20002158) user manual — 4-wire cable types and usage
Page 28 · click to enlarge
Page 28

4-wire cable shielding Figure 3-2:

Gland supplier Micro Motion

type Braided (armored cable)

NPT

Trim 7 mm from the shielded

Apply the Heat Shrink 1. Slide the shielded heat shrink over the drain wires. Ensure that the wires are completely covered. 2. Apply heat (250 °F or 120 °C) to shrink the tubing. Do not burn the cable. 3. Position the clamping insert so the interior end is flush with the braid of the heat shrink.

Emerson Thermostat (20002158) — NPT diagram (page 28)
Fig. 29

Terminate the shield and drain wires in the

Assemble the Gland 1. Fold the shield or braid back over the clamping insert and 1/8 inch (3 mm) past the O-ring. 2. Install the gland body into the conduit opening on the core processor housing. 3. Insert the wires through gland body and tighten the gland nut onto the gland body.

Assemble the gland according to vendor

Emerson Thermostat (20002158) — NPT diagram (page 28)
Fig. 30

4-wire cable types and usage

Micro Motion offers two types of 4-wire cable: shielded and armored. Both types contain shield drain wires.

The 4-wire cable supplied by Micro Motion consists of one pair of red and black 18 AWG (0.75 mm2) wires for the VDC connection, and one pair of white and green 22 AWG (0.35 mm2) wires for the RS-485 connection.

User-supplied 4-wire cable must meet the following requirements:

Show 18 more labels / short text on this page
Wiring · From the preparation · procedure · User-supplied · cable gland · cable gland · Cable shield · Foil (shielded cable) · Gland type M20 · heat shrink · Trim · Shielded heat shrink · After heat applied · gland · instructions · Shield folded back · Gland body · Done
Page 29 of the Emerson Thermostat (20002158) user manual — 3.2.2 Connect the wires to the core processor terminals
Page 29 · click to enlarge
Page 29
  • Twisted pair construction.
  • Applicable hazardous area requirements, if the core processor is installed in a hazardous area.
  • Wire gauge appropriate for the cable length between the core processor and the transmitter.

Wire gauge Table 3-2:

Wire gauge Maximum cable length

VDC 22 AWG (0.35 mm2) 300 ft (90 m)

VDC 20 AWG (0.5 mm2) 500 ft (150 m)

VDC 18 AWG (0.8 mm2) 1000 ft (300 m)

RS-485 22 AWG (0.35 mm2) or larger 1000 ft (300 m)

3.2.2 Connect the wires to the core processor terminals

After the 4-wire cable has been prepared and shielded (if required), connect the individual wires of the 4-wire cable to the terminals on the core processor.

Wiring

Page 30 of the Emerson Thermostat (20002158) user manual — 3.3 Connect 9-wire cable
Page 30 · click to enlarge
Page 30

Core processor terminals Figure 3-3:

From Step 1 or 2

Enhanced core processor Standard core processor

Connect the wires to the core processor terminals: Red wire > Terminal 1 (Power supply +) Black wire > Terminal 2 (Power supply –) White wire > Terminal 3 (RS-485/A) Green wire > Terminal 4 (RS-485/B)

Connect the wires to the core processor terminals: Red wire > Terminal 1 (Power supply +) Black wire > Terminal 2 (Power supply –) White wire > Terminal 3 (RS-485/A) Green wire > Terminal 4 (RS-485/B)

Reinstall and tighten the core processor cover

  • 1. Reinstall the core processor cover. 2. Torque cover screws to:
  • 10–13 in-lbs (1,13–1,47 N-m) for aluminum housing
  • Minimum 19 in-lbs (2,1 N-m) for stainless steel housing

Connect the wires to the transmitter

terminals (see the transmitter manual)

3.3 Connect 9-wire cable

1. Prepare and install the cable according to the instructions in the Micro Motion 9Wire Flowmeter Cable Preparation and Installation Guide .
2. Insert the stripped ends of the individual wires into the terminal blocks. Ensure that no bare wires remain exposed.

Wiring · Core processor · type

Page 31 of the Emerson Thermostat (20002158) user manual — Connect 9-wire cable
Page 31 · click to enlarge
Page 31
3. Match the wires color for color. For wiring at the transmitter or remote core processor, refer to the transmitter documentation.
4. Tighten the screws to hold the wires in place.
5. Ensure integrity of gaskets, then tightly close and seal the junction box cover and all housing covers on the transmitter or core processor.
6. Refer to the transmitter installation manual for signal and power wiring instructions.

Wiring

Page 32 of the Emerson Thermostat (20002158) user manual — Procedure
Page 32 · click to enlarge
Page 32

4 Grounding

The meter must be grounded according to the standards that are applicable at the site. The customer is responsible for knowing and complying with all applicable standards.

Prerequisites

Micro Motion suggests the following guides for grounding practices:

  • In Europe, IEC 79-14 is applicable to most installations, in particular Sections 12.2.2.3 and 12.2.2.4.
  • In the U.S.A. and Canada, ISA 12.06.01 Part 1 provides examples with associated applications and requirements.

If no external standards are applicable, follow these guidelines to ground the sensor:

  • Use copper wire, 14 AWG (2,0 mm 2 ) or larger wire size.
  • Keep all ground leads as short as possible, less than 1 Ω impedance.
  • Connect ground leads directly to earth, or follow plant standards.

CAUTION!

Ground the flowmeter to earth, or follow ground network requirements for the facility. Improper grounding can cause measurement error.

Procedure

Check the joints in the pipeline.

- If the joints in the pipeline are ground-bonded, the sensor is automatically grounded and no further action is necessary (unless required by local code).

- If the joints in the pipeline are not grounded, connect a ground wire to the grounding screw located on the sensor electronics.

Grounding

Page 33 of the Emerson Thermostat (20002158) user manual — Procedure
Page 33 · click to enlarge
Page 33

5 Supplementary information

Topics covered in this chapter:

  • Purge the sensor case
  • About rupture disks

5.1 Purge the sensor case

If the sensor has purge fittings, they should remain sealed at all times. After a purge plug has been removed, the sensor case should be purged with argon or nitrogen and resealed.

Purging the case protects internal components. The sensor is purged of all oxygen and sealed at the factory. If the purge plugs are never removed, it is not necessary to purge or re-seal the sensor. For more information, contact Micro Motion Customer Service.

If a purge plug is removed from the sensor case, it will be necessary to repurge the case.

CAUTION!

Take all necessary precautions when removing purge plugs. Removing a purge plug compromises the secondary containment of the sensor and could expose the user to process fluid.

CAUTION!

Improper pressurization of the sensor case could result in personal injury. Removing a purge plug will require the sensor case to be repurged with a dry inert gas. Follow all instructions provided in the case purging procedure.

Prerequisites

Make sure the following are available before beginning the purge procedure:

  • Teflon ® tape
  • Argon or nitrogen gas sufficient to purge the sensor case

Procedure

1. Shut down the process, or set control devices for manual operation. Before performing the case purging procedure, shut down the process or set the control devices for manual operation. Performing the purge procedure while the flowmeter is operating could affect measurement accuracy, resulting in inaccurate flow signals.
2. Remove both purge plugs from the sensor case. If purge lines are being used, open the valve in the purge lines.
3. Prepare the purge plugs for reinstallation by wrapping them with 3–5 turns of Teflon tape.

Supplementary information

Page 34 of the Emerson Thermostat (20002158) user manual — Purge time Table 5-1:
Page 34 · click to enlarge
Page 34
4. Connect the supply of nitrogen or argon gas to the inlet purge connection or open inlet purge line. Leave the outlet connection open.
  • Exercise caution to avoid introducing dirt, moisture, rust, or other contaminants into the sensor case.
  • If the purge gas is heavier than air (such as argon), locate the inlet lower than the outlet, so that the purge gas will displace air from bottom to top.
  • If the purge gas is lighter than air (such as nitrogen), locate the inlet higher than the outlet, so that the purge gas will displace air from top to bottom.
5. Make sure that there is a tight seal between the inlet connection and sensor case, so that air cannot be drawn by suction into the case or purge line during the purging process.
6. Run purge gas through the sensor.

The purge time is the amount of time required for full exchange of atmosphere to inert gas. The larger the line size, the greater amount of time is required to purge the case. If purge lines are being used, increase the purge time to fill the additional volume of the purge line.

Purge time Table 5-1:

Emerson Thermostat (20002158) — Purge time Table 5-1: diagram (page 34)
Fig. 31 · Purge time Table 5-1:

Sensor model Purge rate, in ft3/hr (l/h) Time, in minutes

Supplementary information

Page 35 of the Emerson Thermostat (20002158) user manual — CAUTION!
Page 35 · click to enlarge
Page 35

Purge time (continued) Table 5-1:

Sensor model Purge rate, in ft3/hr (l/h) Time, in minutes

CMFS100 20 (566) 6

CMFS150 20 (566) 6

7. At the appropriate time, shut off the gas supply, then immediately seal the purge outlet and inlet connections with the purge plugs.
8. Make sure that the purge fitting seals are tight so that air cannot be drawn by suction into the sensor case.

5.2 About rupture disks

Rupture disks are meant to vent process fluid from the sensor case in the event of a flow tube rupture. Some users connect a pipeline to the rupture disk to help contain escaping process fluid. For more information about rupture disks, contact Micro Motion Customer Service.

If the sensor has rupture disks, they are installed in the sensor purge fitting openings. The rupture disks should remain installed at all times. If you remove a rupture disk from the sensor case, it will be necessary to re-purge the case.

CAUTION!

Stay clear of the rupture disk pressure relief area. High-pressure fluid escaping from the sensor can cause severe injury or death.

Supplementary information

Page 36 of the Emerson Thermostat (20002158) user manual — *20002158* 20002158 Rev DI
Page 36 · click to enlarge
Page 36

*20002158* 20002158 Rev DI

Micro Motion Inc. USA Worldwide Headquarters 7070 Winchester Circle Boulder, Colorado 80301 T +1 303-527-5200 T +1 800-522-6277 F +1 303-530-8459 www.micromotion.com

Micro Motion Europe Emerson Process Management Neonstraat 1 6718 WX Ede The Netherlands T +31 (0) 318 495 555 F +31 (0) 318 495 556 www.micromotion.nl

Micro Motion Asia Emerson Process Management 1 Pandan Crescent Singapore 128461 Republic of Singapore T +65 6777-8211 F +65 6770-8003

Micro Motion United Kingdom Emerson Process Management Limited Horsfield Way Bredbury Industrial Estate Stockport SK6 2SU U.K. T +44 0870 240 1978 F +44 0800 966 181

Micro Motion Japan Emerson Process Management 1-2-5, Higashi Shinagawa Shinagawa-ku Tokyo 140-0002 Japan T +81 3 5769-6803 F +81 3 5769-6844

©2014 Micro Motion, Inc. All rights reserved.

The Emerson logo is a trademark and service mark of Emerson Electric Co. Micro Motion, ELITE, ProLink, MVD and MVD Direct Connect marks are marks of one of the Emerson Process Management family of companies. All other marks are property of their respective owners.

Sources

References & links

Pages referenced in this manual — manufacturer, support and downloads.

Ask about the Emerson Thermostat (20002158)

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Emerson Thermostat (20002158)

Be the first to ask — other owners or our team will answer.

Original document

Emerson Thermostat (20002158) manual PDF

36 pages·3.3 MB·Thermostat·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…