Agilent Technologies PSA Spectrum Analyzers
This manual provides documentation for the following instruments:
E4440A (3 Hz – 26.5 GHz) E4443A (3 Hz – 6.7 GHz) E4445A (3 Hz – 13.2 GHz) E4446A (3 Hz – 44.0 GHz) E4448A (3 Hz – 50.0 GHz)

Printed in USA December 2002
© Copyright 2001 Agilent Technologies

The information contained in this document is subject to change without notice.
Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
Where to Find the Latest Information
Documentation is updated periodically. For the latest information about Agilent PSA spectrum analyzers, including firmware upgrades and application information, see: http//www.agilent.com/find/psa



:CALCulate:LLINe:ALL:DELete 122
:CALCulate:LLINe:CMODe FIXed|RELative 121
:CALCulate:LLINe:CMODe? 121
:CALCulate:LLINe:CONTrol:DOMain FREQuency|TIME 120
:CALCulate:LLINe:CONTrol:DOMain? 120
:CALCulate:LLINe[1]|2:AMPLitude:INTerpolate:TYPE LOGarithmic|LINear 119
:CALCulate:LLINe[1]|2:AMPLitude:INTerpolate:TYPE? 119
:CALCulate:LLINe[1]|2:CONTrol:INTerpolate:TYPE LOGarithmic|LINear 119
:CALCulate:LLINe[1]|2:CONTrol:INTerpolate:TYPE? 119
:CALCulate:LLINe[1]|2:DATA , , {,,,} 109
:CALCulate:LLINe[1]|2:DATA:MERGe , , {,,,} 109
:CALCulate:LLINe[1]|2:DATA? 109
:CALCulate:LLINe[1]|2:DELete 116
:CALCulate:LLINe[1]|2:DISPlay 111
:CALCulate:LLINe[1]|2:DISPlay? 111
:CALCulate:LLINe[1]|2:FAIL? 111
:CALCulate:LLINe[1]|2:MARGin 113
:CALCulate:LLINe[1]|2:MARGin:STATe OFF|ON|0|1 112
:CALCulate:LLINe[1]|2:MARGin:STATe? 112
:CALCulate:LLINe[1]|2:MARGin? 113
:CALCulate:LLINe[1]|2:STATe OFF|ON|0|1 111
:CALCulate:LLINe[1]|2:STATe? 111
:CALCulate:LLINe[1]|2:TYPE UPPer|LOWer 110
:CALCulate:LLINe[1]|2:TYPE? 110
:CALCulate:MARKer:AOFF 189
:CALCulate:MARKer:FCOunt:GATetime 196
:CALCulate:MARKer:FCOunt:GATetime:AUTO OFF|ON|0|1 196
:CALCulate:MARKer:FCOunt:GATetime:AUTO? 196
:CALCulate:MARKer:FCOunt:GATetime? 196
:CALCulate:MARKer:PEAK[1]|2|3|4:SEARch:MODE PARameter|MAXimum 218
List of Commands


:CALCulate:MARKer[1]|2|3|4:X:CENTer? 185
:CALCulate:MARKer[1]|2|3|4:X:POSition:CENTer 184
:CALCulate:MARKer[1]|2|3|4:X:POSition:CENTer? 184
:CALCulate:MARKer[1]|2|3|4:X:POSition:SPAN 185
:CALCulate:MARKer[1]|2|3|4:X:POSition:SPAN? 185
:CALCulate:MARKer[1]|2|3|4:X:POSition:STARt 182
:CALCulate:MARKer[1]|2|3|4:X:POSition:STARt? 182
:CALCulate:MARKer[1]|2|3|4:X:POSition:STOP 182
:CALCulate:MARKer[1]|2|3|4:X:POSition:STOP? 182
:CALCulate:MARKer[1]|2|3|4:X:READout FREQuency|TIME|ITIMe|PERiod 186
:CALCulate:MARKer[1]|2|3|4:X:READout? 187
:CALCulate:MARKer[1]|2|3|4:X:SPAN 185
:CALCulate:MARKer[1]|2|3|4:X:SPAN? 185
:CALCulate:MARKer[1]|2|3|4:X? 179
:CALCulate:MARKer[1]|2|3|4:Y? 179
:CALCulate:MARKer[1]|2|3|4[:SET]:CENTer 197
:CALCulate:MARKer[1]|2|3|4[:SET]:DELTa:CENTer 199
:CALCulate:MARKer[1]|2|3|4[:SET]:DELTa:SPAN 199
:CALCulate:MARKer[1]|2|3|4[:SET]:RLEVel 200
:CALCulate:MARKer[1]|2|3|4[:SET]:STARt 197
:CALCulate:MARKer[1]|2|3|4[:SET]:STEP 197
:CALCulate:MARKer[1]|2|3|4[:SET]:STOP 199
:CALCulate:MCPower:OFFSet:LIST:LIMit:NEGative[:UPPer]:DATA ,, 440
:CALCulate:MCPower:OFFSet:LIST:LIMit:NEGative[:UPPer]:DATA? 440
:CALCulate:MCPower:OFFSet:LIST:LIMit:POSitive[:UPPer]:DATA ,, 440
:CALCulate:MCPower:OFFSet:LIST:LIMit:POSitive[:UPPer]:DATA? 440
:CALCulate:PSTatistic:MARKer[1]|2|3|4:MAXimum 397
:CALCulate:PSTatistic:MARKer[1]|2|3|4:MINimum 397
:CALCulate:PSTatistic:MARKer[1]|2|3|4:MODE NORMal|DELTa|OFF 391
:CALCulate:PSTatistic:MARKer[1]|2|3|4:MODE? 391
:CALCulate:PSTatistic:MARKer[1]|2|3|4:TRACe MEASured|GAUSsian|REFerence 394
List of Commands


:CALibration:IF? 259
:CALibration:RF 258
:CALibration:RF? 258
:CALibration:TCORrections AUTO|ON|OFF 262
:CALibration[:ALL] 257
:CALibration[:ALL]? 257
:CONFigure:ACPower 306
:CONFigure:BPOWer 320
:CONFigure:CHPower 303
:CONFigure:HARMonics 316
:CONFigure:MCP 310
:CONFigure:OBW 304
:CONFigure:PSTatistic 314
:CONFigure:SANalyzer 300
:CONFigure:SEMask 324
:CONFigure:SPURious 322
:CONFigure:TOIN 321
:CONFigure? 328
:COUPle ALL|NONE 72
:COUPle? 72
:DISPlay:AFUNction:POSition BOTtom 108
:DISPlay:AFUNction:POSition BOTtom|CENTer|TOP 106
:DISPlay:AFUNction:POSition CENTer 107
:DISPlay:AFUNction:POSition TOP 107
:DISPlay:AFUNction:POSition? 106
:DISPlay:ANNotation:CLOCk:DATE:FORMat MDY|DMY 253
:DISPlay:ANNotation:CLOCk:DATE:FORMat? 253
:DISPlay:ANNotation:CLOCk[:STATe] ON|OFF 253
:DISPlay:ANNotation:CLOCk[:STATe]? 253
:DISPlay:ANNotation:TITLe:DATA 122
:DISPlay:ANNotation:TITLe:DATA? 122
List of Commands


:FETCh:CHPower:DENSity? 303
:FETCh:CHPower? 303
:FETCh:HARMonics:AMPLitude:ALL? 316
:FETCh:HARMonics:AMPLitude[n]? 316
:FETCh:HARMonics:FREQuency:ALL? 316
:FETCh:HARMonics:FREQuency[n]? 317
:FETCh:HARMonics:FUNDamental? 317
:FETCh:HARMonics[:DISTortion]? 316
:FETCh:HARMonics[n]? 317
:FETCh:MCP[n]? 310
:FETCh:MCPower? 310
:FETCh:MCPower? 311
:FETCh:OBW:FERRor? 304
:FETCh:OBW:OBWidth? 304
:FETCh:OBW:XDB? 304
:FETCh:OBW? 304
:FETCh:PSTatistic[n]? 314
:FETCh:SEMask[n]? 324
:FETCh:SPURious[n]? 322
:FETCh:TOIN:IP2? 321
:FETCh:TOIN:IP3? 321
:FETCh:TOIN? 321
:FORMat:BORDer NORMal|SWAPped 160
:FORMat:BORDer? 160
:FORMat[:TRACe][:DATA] ASCii|INTeger,32|REAL,32|REAL,64|UINTeger,16 161
:FORMat[:TRACe][:DATA]? 161
:HCOPy:ABORt 227
:HCOPy:DEVice:COLor NO|YES 230
:HCOPy:DEVice:COLor? 230
:HCOPy:DEVice:LANGuage PCL3|PCL5 230
:HCOPy:DEVice:LANGuage? 230
List of Commands


:INSTrument:NSELect? 205
:INSTrument[:SELect] BASIC|CDMA|CDMA1XEV|CDMA2K|EDGEGSM|LINK|NADC|NFIG- URE|PDC|PNOISE|SA|WCDMA 203
:INSTrument[:SELect]? 203
:MEASure:ACPower[n]? 306
:MEASure:BPOWer[n]? 320
:MEASure:CHPower:CHPower? 303
:MEASure:CHPower:DENSity? 303
:MEASure:CHPower? 303
:MEASure:HARMonics:AMPLitude:ALL? 316
:MEASure:HARMonics:AMPLitude[n]? 316
:MEASure:HARMonics:FREQuency:ALL? 316
:MEASure:HARMonics:FREQuency[n]? 317
:MEASure:HARMonics:FUNDamental? 317
:MEASure:HARMonics[:DISTortion]? 316
:MEASure:HARMonics[n]? 317
:MEASure:MCP[n]? 310
:MEASure:MCPower? 310
:MEASure:MCPower? 311
:MEASure:OBW:FERRor? 304
:MEASure:OBW:OBWidth? 304
:MEASure:OBW:XDB? 304
:MEASure:OBW? 304
:MEASure:PSTatistic[n]? 314
:MEASure:SEMask[n]? 324
:MEASure:SPURious[n]? 322
:MEASure:TOIN:IP2? 321
:MEASure:TOIN:IP3? 321
:MEASure:TOIN? 321
:MMEMory:CATalog? 130
:MMEMory:COPY , 151
List of Commands


:READ:OBW:FERRor? 304
:READ:OBW:OBWidth? 304
:READ:OBW:XDB? 304
:READ:OBW? 304
:READ:PSTatistic[n]? 314
:READ:SEMask[n]? 324
:READ:SPURious[n]? 322
:READ:TOIN:IP2? 321
:READ:TOIN:IP3? 321
:READ:TOIN? 321
:STATus:OPERation:CONDition? 571
:STATus:OPERation:ENABle 571
:STATus:OPERation:ENABle? 571
:STATus:OPERation:NTRansition 572
:STATus:OPERation:NTRansition? 572
:STATus:OPERation:PTRansition 572
:STATus:OPERation:PTRansition? 572
:STATus:OPERation[:EVENt]? 572
:STATus:PRESet 573
:STATus:QUEStionable:CALibration:CONDition? 575
:STATus:QUEStionable:CALibration:ENABle 575
:STATus:QUEStionable:CALibration:ENABle? 575
:STATus:QUEStionable:CALibration:NTRansition 576
:STATus:QUEStionable:CALibration:NTRansition? 576
:STATus:QUEStionable:CALibration:PTRansition 576
:STATus:QUEStionable:CALibration:PTRansition? 576
:STATus:QUEStionable:CALibration[:EVENt]? 575
:STATus:QUEStionable:CONDition? 573
:STATus:QUEStionable:ENABle 573
:STATus:QUEStionable:ENABle? 573
:STATus:QUEStionable:FREQuency:CONDition? 576
List of Commands


:STATus:QUEStionable:NTRansition 574
:STATus:QUEStionable:NTRansition? 574
:STATus:QUEStionable:POWer:CONDition? 583
:STATus:QUEStionable:POWer:ENABle 584
:STATus:QUEStionable:POWer:ENABle? 584
:STATus:QUEStionable:POWer:NTRansition 584
:STATus:QUEStionable:POWer:NTRansition? 584
:STATus:QUEStionable:POWer:PTRansition 585
:STATus:QUEStionable:POWer:PTRansition?> 585
:STATus:QUEStionable:POWer[:EVENt]? 584
:STATus:QUEStionable:PTRansition 574
:STATus:QUEStionable:PTRansition? 574
:STATus:QUEStionable:TEMPerature:CONDition? 585
:STATus:QUEStionable:TEMPerature:ENABle 585
:STATus:QUEStionable:TEMPerature:ENABle? 585
:STATus:QUEStionable:TEMPerature:NTRansition 586
:STATus:QUEStionable:TEMPerature:NTRansition? 586
:STATus:QUEStionable:TEMPerature:PTRansition 586
:STATus:QUEStionable:TEMPerature:PTRansition? 586
:STATus:QUEStionable:TEMPerature[:EVENt]? 586
:STATus:QUEStionable[:EVENt]? 574
:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess 263
:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess? 263
:SYSTem:COMMunicate:LAN:SCPI:SICL:ENABle OFF|ON|0|1 267
:SYSTem:COMMunicate:LAN:SCPI:SICL:ENABle? 267
:SYSTem:COMMunicate:LAN:SCPI:SOCKet:ENABle OFF|ON|0|1 266
:SYSTem:COMMunicate:LAN:SCPI:SOCKet:ENABle? 266
:SYSTem:COMMunicate:LAN:SCPI:TELNet:ENABle OFF|ON|0|1 266
:SYSTem:COMMunicate:LAN:SCPI:TELNet:ENABle? 266
:SYSTem:COMMunicate:LAN[:SELF]:IP 263
:SYSTem:COMMunicate:LAN[:SELF]:IP? 263
List of Commands


:TRACe[1]|2|3:MODE? 283
:TRACe[1]|2|3:MODE? 369
:TRACe[1]|2|3:MODE? 379
:TRACe[1]|2|3:MODE? 406
:TRACe[1]|2|3:MODE? 448
:TRIGger[:SEQuence]:DELay 290
:TRIGger[:SEQuence]:DELay:STATe OFF|ON|0|1 290
:TRIGger[:SEQuence]:DELay:STATe? 290
:TRIGger[:SEQuence]:DELay? 290
:TRIGger[:SEQuence]:OFFSet 291
:TRIGger[:SEQuence]:OFFSet:STATe OFF|ON|0|1 291
:TRIGger[:SEQuence]:OFFSet:STATe? 291
:TRIGger[:SEQuence]:OFFSet? 291
:TRIGger[:SEQuence]:SLOPe POSitive|NEGative 290
:TRIGger[:SEQuence]:SLOPe? 290
:TRIGger[:SEQuence]:SOURce IMMediate|VIDeo|LINE|EXTernal[1]|EXTernal2|RFBurst . .287
:TRIGger[:SEQuence]:SOURce? 287
:TRIGger[:SEQuence]:VIDeo:LEVel 288
:TRIGger[:SEQuence]:VIDeo:LEVel:FREQuency 288
:TRIGger[:SEQuence]:VIDeo:LEVel:FREQuency? 288
:TRIGger[:SEQuence]:VIDeo:LEVel? 288
:UNIT:POWer DBM|DBMV|DBUV|V|W 56
:UNIT:POWer? 56
[:SENCe]:CORRection:IMPedance[:INPut][:MAGNitude] 50|75 173
[:SENCe]:CORRection:IMPedance[:INPut][:MAGNitude]? 173
[:SENSe]:ACPower:AVERage:COUNt 351
[:SENSe]:ACPower:AVERage:COUNt? 351
[:SENSe]:ACPower:AVERage:TCONrol EXPonential|REPeat 353
[:SENSe]:ACPower:AVERage:TCONrol? 353
[:SENSe]:ACPower:AVERage[:STATe] OFF|ON|0|1 351
[:SENSe]:ACPower:AVERage[:STATe]? 351
List of Commands


[:SENSe]:AVERage:COUNt? 89
[:SENSe]:AVERage:TYPE RMS|LOG|SCALar 90
[:SENSe]:AVERage:TYPE:AUTO OFF|ON|0|1 90
[:SENSe]:AVERage:TYPE:AUTO? 90
[:SENSe]:AVERage:TYPE? 90
[:SENSe]:AVERage[:STATe] OFF|ON|0|1 89
[:SENSe]:AVERage[:STATe]? 89
[:SENSe]:BANDwidth|BWIDth:VIDeo 86
[:SENSe]:BANDwidth|BWIDth:VIDeo:AUTO OFF|ON|0|1 86
[:SENSe]:BANDwidth|BWIDth:VIDeo:AUTO? 86
[:SENSe]:BANDwidth|BWIDth:VIDeo:RATio 87
[:SENSe]:BANDwidth|BWIDth:VIDeo:RATio:AUTO OFF|ON|0|1 87
[:SENSe]:BANDwidth|BWIDth:VIDeo:RATio:AUTO? 87
[:SENSe]:BANDwidth|BWIDth:VIDeo:RATio? 87
[:SENSe]:BANDwidth|BWIDth:VIDeo? 86
[:SENSe]:BANDwidth|BWIDth[:RESolution] 84
[:SENSe]:BANDwidth|BWIDth[:RESolution]:AUTO OFF|ON|0|1 84
[:SENSe]:BANDwidth|BWIDth[:RESolution]:AUTO? 84
[:SENSe]:BANDwidth|BWIDth[:RESolution]? 84
[:SENSe]:BPOWer:AVERage:COUNt 371
[:SENSe]:BPOWer:AVERage:COUNt? 371
[:SENSe]:BPOWer:AVERage:TCONrol EXPonential|REPeat 372
[:SENSe]:BPOWer:AVERage:TCONrol? 372
[:SENSe]:BPOWer:AVERage:TYPe LPOWer|POWer 372
[:SENSe]:BPOWer:AVERage:TYPe? 373
[:SENSe]:BPOWer:AVERage[:STATe] OFF|ON|0|1 371
[:SENSe]:BPOWer:AVERage[:STATe]? 371
[:SENSe]:BPOWer:BURSt:AUTO OFF|ON|0|1 375
[:SENSe]:BPOWer:BURSt:AUTO? 375
[:SENSe]:BPOWer:BURSt:WIDTh 375
[:SENSe]:BPOWer:BURSt:WIDTh? 375
List of Commands


[:SENSe]:CORRection:OFFSet[:MAGNitude]? 68
[:SENSe]:DETector:AUTO OFF|ON|0|1 98
[:SENSe]:DETector:AUTO? 98
[:SENSe]:DETector[:FUNCtion] AVERage|NEGative|NORMal|POSitive|SAMPle|RMS 96
[:SENSe]:DETector[:FUNCtion]? 96
[:SENSe]:FEED RF|AREFerence 171
[:SENSe]:FEED? 171
[:SENSe]:FREQuency:CENTer |UP|DOWN 164
[:SENSe]:FREQuency:CENTer:STEP:AUTO OFF|ON|0|1 167
[:SENSe]:FREQuency:CENTer:STEP:AUTO? 167
[:SENSe]:FREQuency:CENTer:STEP[:INCRement] 167
[:SENSe]:FREQuency:CENTer:STEP[:INCRement]? 167
[:SENSe]:FREQuency:CENTer? 164
[:SENSe]:FREQuency:OFFSet 168
[:SENSe]:FREQuency:OFFSet? 168
[:SENSe]:FREQuency:SPAN 238
[:SENSe]:FREQuency:SPAN:BANDwidth[:RESolution]:RATio 93
[:SENSe]:FREQuency:SPAN:BANDwidth[:RESolution]:RATio:AUTO OFF|ON|0|1 93
[:SENSe]:FREQuency:SPAN:BANDwidth[:RESolution]:RATio:AUTO? 93
[:SENSe]:FREQuency:SPAN:BANDwidth[:RESolution]:RATio? 93
[:SENSe]:FREQuency:SPAN:FULL 238
[:SENSe]:FREQuency:SPAN:PREVious 239
[:SENSe]:FREQuency:SPAN? 238
[:SENSe]:FREQuency:STARt 165
[:SENSe]:FREQuency:STARt? 165
[:SENSe]:FREQuency:STOP 166
[:SENSe]:FREQuency:STOP? 166
[:SENSe]:FREQuency:SYNThesis 1|2|3 76
[:SENSe]:FREQuency:SYNThesis:AUTO OFF|ON|0|1 76
[:SENSe]:FREQuency:SYNThesis:AUTO? 76
[:SENSe]:FREQuency:SYNThesis? 76
List of Commands


[:SENSe]:HARMonics:SWEeptime:STATe? 412
[:SENSe]:HARMonics:SWEeptime? 412
[:SENSe]:MARKer[1]|2|3|4:X:POSition 179
[:SENSe]:MARKer[1]|2|3|4:X:POSition? 179
[:SENSe]:MCPower:AVERage:COUNt 429
[:SENSe]:MCPower:AVERage:COUNt? 429
[:SENSe]:MCPower:AVERage:TCONrol EXPonential|REPeat 430
[:SENSe]:MCPower:AVERage:TCONrol? 430
[:SENSe]:MCPower:AVERage[:STATe] OFF|ON|0|1 429
[:SENSe]:MCPower:AVERage[:STATe]? 429
[:SENSe]:MCPower:CARRier:AUTO[:STATe] OFF|ON|0|1 442
[:SENSe]:MCPower:CARRier:AUTO[:STATe]? 442
[:SENSe]:MCPower:CARRier:COUNt 431
[:SENSe]:MCPower:CARRier:COUNt? 431
[:SENSe]:MCPower:CARRier:LIST:BANDwidth|BWIDth:[INTegration] 438
[:SENSe]:MCPower:CARRier:LIST:BANDwidth|BWIDth:[INTegration]? 438
[:SENSe]:MCPower:CARRier:LIST:PPResent YES|NO 435
[:SENSe]:MCPower:CARRier:LIST:PPResent? 435
[:SENSe]:MCPower:CARRier:LIST:WIDTh 437
[:SENSe]:MCPower:CARRier:LIST:WIDTh? 437
[:SENSe]:MCPower:CARRier[:POWer] 442
[:SENSe]:MCPower:CARRier[:POWer]? 442
[:SENSe]:MCPower:CORRection:NOISe[:AUTO] OFF|ON|0|1 445
[:SENSe]:MCPower:CORRection:NOISe[:AUTO]? 445
[:SENSe]:MCPower:FILTer[:RRC]:ALPHA 445
[:SENSe]:MCPower:FILTer[:RRC]:ALPHA? 445
[:SENSe]:MCPower:FILTer[:RRC][:STATe] OFF|ON|0|1 444
[:SENSe]:MCPower:LIMit[:STATe] OFF|ON|0|1 444
[:SENSe]:MCPower:LIMit[:STATe]? 444
[:SENSe]:MCPower:METHod IBW|RBW 442
[:SENSe]:MCPower:METHod? 442
List of Commands


[:SENSe]:POWer[:RF]:GAIN[:STATe] OFF|ON|0|1 59
[:SENSe]:POWer[:RF]:GAIN[:STATe]? 59
[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer] 69
[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer]? 69
[:SENSe]:POWer[:RF]:MMW:PADJust 55
[:SENSe]:POWer[:RF]:MMW:PADJust? 56
[:SENSe]:POWer[:RF]:MW:PADJust 55
[:SENSe]:POWer[:RF]:MW:PADJust? 55
[:SENSe]:POWer[:RF]:PCENter 54
[:SENSe]:POWer[:RF]:RANGe:AUTO ONCE 383
[:SENSe]:PSTatistic:BANDwidth|BWIDth 381
[:SENSe]:PSTatistic:BANDwidth|BWIDth? 381
[:SENSe]:PSTatistic:COUNts 382
[:SENSe]:PSTatistic:COUNts? 382
[:SENSe]:PSTatistic:SWEep:TIME 383
[:SENSe]:PSTatistic:SWEep:TIME? 383
[:SENSe]:RADio:STANdard:BAND:CLASs BC0|BC1 342
[:SENSe]:RADio:STANdard:BAND:CLASs? 342
[:SENSe]:RADio:STANdard:DEVice BTS|MS 341
[:SENSe]:RADio:STANdard:DEVice? 341
[:SENSe]:RADio:STANdard:EAMeas 344
[:SENSe]:RADio:STANdard:EAMeas 344
[:SENSe]:RADio:STANdard:PACKet DH1|DH3|DH5 342
[:SENSe]:RADio:STANdard:PACKet? 342
[:SENSe]:RADio:STANdard:RPARams[:STATe]? 344
[:SENSe]:RADio:STANdard:RPARams[:STATe]ON|Off|1|0 344
[:SENSe]:RADio:STANdard:SBWidth 341
[:SENSe]:RADio:STANdard:SBWidth? 341
[:SENSe]:RADio:STANdard[:SE- Lect]NONE|IS95a|JSTD|IS95C|GSM|W3GPP|C2000MC1|C2000DS|C2000MC3|NADC|PDC |BLUEtooth|TETRa |WL802DOT11A|WL802DOT11B|WL802DOT11G|HIPERLAN2||DVBT- GPN|DVBTIPN 333
List of Commands


[:SENSe]:SEMask:OFFSet[n]:LIST:BANDwidth|BWIDth[:RESolution]:AUTO OFF|ON|0|1 . .466
[:SENSe]:SEMask:OFFSet[n]:LIST:BANDwidth|BWIDth[:RESolution]:AUTO? 466
[:SENSe]:SEMask:OFFSet[n]:LIST:BANDwidth|BWIDth[:RESolution]? 466
[:SENSe]:SEMask:OFFSet[n]:LIST:FREQuency:STARt , ... [] 464
[:SENSe]:SEMask:OFFSet[n]:LIST:FREQuency:STARt? 464
[:SENSe]:SEMask:OFFSet[n]:LIST:FREQuency:STOP ... [] 465
[:SENSe]:SEMask:OFFSet[n]:LIST:FREQuency:STOP? 465
[:SENSe]:SEMask:OFFSet[n]:LIST:STARt:ABSolute ,...[] (up to five values) 469
[:SENSe]:SEMask:OFFSet[n]:LIST:STARt:ABSolute? 469
[:SENSe]:SEMask:OFFSet[n]:LIST:STARt:RCARrier ,... [ ] (up to five values) 471
[:SENSe]:SEMask:OFFSet[n]:LIST:STARt:RCARrier? 471
[:SENSe]:SEMask:OFFSet[n]:LIST:STATe OFF|ON|0| 464
[:SENSe]:SEMask:OFFSet[n]:LIST:STATe? 464
[:SENSe]:SEMask:OFFSet[n]:LIST:STOP:ABSolute ,...[] (up to five values) 470
[:SENSe]:SEMask:OFFSet[n]:LIST:STOP:ABSolute:COUPle OFF|ON|0|1 470
[:SENSe]:SEMask:OFFSet[n]:LIST:STOP:ABSolute:COUPle? 470
[:SENSe]:SEMask:OFFSet[n]:LIST:STOP:ABSolute? 470
[:SENSe]:SEMask:OFFSet[n]:LIST:STOP:RCARrier ,[] 472
[:SENSe]:SEMask:OFFSet[n]:LIST:STOP:RCARrier:COUPle OFF|ON|0|1 472
[:SENSe]:SEMask:OFFSet[n]:LIST:STOP:RCARrier:COUPle? 472
[:SENSe]:SEMask:OFFSet[n]:LIST:STOP:RCARrier? 472
[:SENSe]:SEMask:OFFSet[n]:LIST:SWEeptime ... [] 465
[:SENSe]:SEMask:OFFSet[n]:LIST:SWEeptime:AUTO OFF|ON|0|1 466
[:SENSe]:SEMask:OFFSet[n]:LIST:SWEeptime:AUTO? 466
[:SENSe]:SEMask:OFFSet[n]:LIST:SWEeptime? 466
[:SENSe]:SEMask:OFFSet[n]:LIST:TEST ABSolute|AND|OR|RELative,ABSo- lute|AND|OR|RELative,ABSolute|AND|OR|RELative,ABSolute|AND|OR|RELative 474
[:SENSe]:SEMask:OFFSet[n]:LIST:TEST? 474
[:SENSe]:SEMask:SWEeptime 460
[:SENSe]:SEMask:SWEeptime:AUTO OFF|ON|0|1 460
[:SENSe]:SEMask:SWEeptime:AUTO? 460
List of Commands


[:SENSe]:SPURious[:RANGe][:LIST]:PEAK:EXCursion 496
[:SENSe]:SPURious[:RANGe][:LIST]:PEAK:EXCursion? 496
[:SENSe]:SPURious[:RANGe][:LIST]:PEAK:THReshold 496
[:SENSe]:SPURious[:RANGe][:LIST]:PEAK:THReshold? 497
[:SENSe]:SPURious[:RANGe][:LIST]:STATe OFF|ON|0|1 488
[:SENSe]:SPURious[:RANGe][:LIST]:STATe? 488
[:SENSe]:SPURious[:RANGe][:LIST]:SWEep:TIME: , ,... (up to 20 sweeptimes) . .492
[:SENSe]:SPURious[:RANGe][:LIST]:SWEep:TIME:AUTO OFF|ON|0|1, OFF|ON|0|1 492
[:SENSe]:SPURious[:RANGe][:LIST]:SWEep:TIME:AUTO? 492
[:SENSe]:SPURious[:RANGe][:LIST]:SWEep:TIME? 492
[:SENSe]:SWEep:FFT:SPAN:RATio 76
[:SENSe]:SWEep:FFT:SPAN:RATio? 76
[:SENSe]:SWEep:POINts 245
[:SENSe]:SWEep:POINts? 245
[:SENSe]:SWEep:TIME 242
[:SENSe]:SWEep:TIME:AUTO OFF|ON|0|1 242
[:SENSe]:SWEep:TIME:AUTO:RULes NORMal|ACCuracy 243
[:SENSe]:SWEep:TIME:AUTO:RULes? 243
[:SENSe]:SWEep:TIME:AUTO? 242
[:SENSe]:SWEep:TIME? 242
[:SENSe]:SWEep:TYPE AUTO|FFT|SWEep 72
[:SENSe]:SWEep:TYPE:AUTO:RULes SPEed|DRANge 73
[:SENSe]:SWEep:TYPE:AUTO:RULes? 73
[:SENSe]:SWEep:TYPE? 72
[:SENSe]:TOIN:AVERage:COUNt 423
[:SENSe]:TOIN:AVERage:COUNt? 423
[:SENSe]:TOIN:AVERage:TCONrol EXPonential|REPeat 425
[:SENSe]:TOIN:AVERage:TCONrol? 425
[:SENSe]:TOIN:AVERage[:STATe] OFF|ON|0|1 423
[:SENSe]:TOIN:AVERage[:STATe]? 423
[:SENSe]:TOIN:FREQuency:MIXer:RANGe:AUTO OFF|ON|0|1 426
List of Commands


List of Commands

This book provides you with descriptions and programming information for your analyzer.
Using This Document

There are many terms used throughout this book, for example “active function block,” that are explained in detail in the Getting Started guide. It is recommended that you read the Getting Started guide first.
Using This Document · Book Organization · Using This Document

The following terms are used to describe each key. Note that a key description may not use all the terms.
State Saved: Indicates what happens to a particular function when the instrument state is saved (either to floppy disk or the internal c:\ drive). It also indicates whether the current settings of the function are maintained if the instrument is powered on or preset using Power On Last State or User Preset.
Dependencies/ Couplings: Describes dependencies or interactions to other functions or settings in the analyzer.
Factory Preset: Describes the function settings after a Factory Preset.
Default Terminator: Indicates the units that will be attached to the numerical value that you have entered. This default will be used from the front panel, when you terminate your entry by pressing the Enter key, rather then selecting a units key. This default will be used remotely when you send the command without specifying any units after your value(s).
Range: Describes the range of the smallest to largest values to which the function can be set. If you try to set a value below the minimum value, the analyzer defaults to the minimum value. If you try to set a value above the maximum value, the analyzer defaults to the maximum value.
History: Describes the firmware revision history. Only applies after first firmware release.
Remote Command: Shows the syntax requirements for each SCPI command.
Example: Provides command examples using the indicated remote command syntax.
Using This Document · Book Organization · Using This Document

Using This Document · Book Organization · Using This Document

This chapter provides key descriptions and programming information for the front-panel key functions of your analyzer starting with the letters A through L. The front-panel functions are listed alphabetically and are described with their associated menu keys. The lower-level menu keys are arranged and described as they appear in your analyzer.


AMPLITUDE / Y Scale
AMPLITUDE / Y Scale
Activates the Reference Level function and displays the Amplitude menu keys. These functions control how data on the vertical (Y) axis is displayed and corrected, and control instrument settings that affect the vertical axis.
Enables you to adjust the absolute amplitude represented by the top graticule line on the display (the reference level). Ref in the upper left corner of the display, indicates the current value. The Amplitude, Y Axis Units setting determines the Reference Level units. To change the reference level, use the front-panel step keys, knob, or numeric keypad.
Dependencies/ Couplings: If you reduce the Attenuation setting, the analyzer may have to lower the Reference Level to maintain the proper level at the top of the screen. If you then increase Attenuation, the Reference Level does not increase to its previous value.
When the input attenuator is auto-coupled, Attenuation (Auto), its setting may be affected by changes in the reference level setting. See “Attenuation” on page 50.
Reference level is affected by: Attenuation, Preamp, Ext Amp Gain, Reference level offset, Max Mixer Level, Y-Axis Units.
Range: Determined by the settings of the input attenuator, reference level offset, external amplitude gain, and whether the preamp (Option 1DS) is on or off.
Examples:
−170 dBm to 30 dBm with zero reference level offset
−180 dBm to 20 dBm with 10 dB ext amp gain
−160 dBm to 40 dBm with 10 dB reference level offset
−170 dBm to 0 dBm with preamp on (Option 1DS)
:DISPlay:WINDow[1]:TRACe:Y:[SCALe]:RLEVel
:DISPlay:WINDow[1]:TRACe:Y:[SCALe]:RLEVel?
Sets the reference level to 20 dBm, which displays in the current Y-Axis Units. For example, if the Y-Axis Units are dBµV, then 127 dBµV will be displayed.

AMPLITUDE / Y Scale
Allows you to adjust the input attenuation. Press Atten Step to set the attenuation step so that attenuation will change in 2 dB or 10 dB increments. The analyzer input attenuator reduces the power level of the input signal delivered to the input mixer. If set manually, the attenuator is recoupled when Attenuation (Auto) is selected. To enter a value below 6 dB, you must use the front-panel numeric keypad.
Attenuation is coupled to Reference Level, so adjusting the Reference Level may change the Attenuation. The analyzer selects an Attenuation setting that is as small as possible while keeping the Ref Level at or below the Max Mixer Lvl setting. The current value is indicated by Atten at the top of the display. A # appears in front of Atten when Attenuation (Man) is selected.
To prevent signal compression, keep the power at the input mixer below 0 dBm (10 MHz - 200 MHz), below 3 dBm (200 MHz - 6.6 GHz), and below –2 dBm (6.6 GHz - 50.0 GHz). With the attenuator set to Auto, a signal at or below the reference level results in a mixer level at or below −10 dBm.
Resolution/Rounding/ Truncation: The attenuation is resolved to 2 dB increments. Value setting a logarithm: if value is at least 0.5 dB over a value, then the next higher value is selected. Therefore, 10.4 selects 10, while 10.5 selects 12.
[:SENSe]:POWer[:RF]:ATTenuation
[:SENSe]:POWer[:RF]:ATTenuation?
[:SENSe]:POWer[:RF]:ATTenuation:AUTO OFF|ON|0|1
[:SENSe]:POWer[:RF]:ATTenuation:AUTO?
Remote Command Notes: The Reference Level setting may be affected when the Attenuation is changed. See Ref Level.
POW:ATT:AUTO ON

AMPLITUDE / Y Scale

AMPLITUDE / Y Scale
Sets the units per vertical graticule division on the display. This function is only available when Scale Type (Log) is selected and the vertical scale is power, or Demod View is on and the vertical scale is hertz. When Scale Type (Lin) is selected, Scale/Div is greyed out.
1 kHz to 240 kHz (When Demod View (On) is selected.)
:DISPlay:WINDow[1]:TRACe:Y:[SCALe]:PDIVision
:DISPlay:WINDow[1]:TRACe:Y:[SCALe]:PDIVision?
Allows you to choose a linear or logarithmic vertical scale for the display and for remote data readout.
The scale type for display and remote data readout may be different from the scale used for averaging processes. For information on the scale used for averaging process, “Avg/VBW Type” on page 89.
When Scale Type (Log) is selected, the vertical graticule divisions are scaled in logarithmic units. The top line of the graticule is the Reference Level and use the scaling per division, Scale/Div to assign values to the other locations on the graticule.
When Scale Type (Lin) is selected, the vertical graticule divisions are linearly scaled with the reference level value at the top of the display and zero volts at the bottom. Each vertical division of the graticule represents one-tenth of the Reference Level.
The Y Axis Units used for each type of display are set by pressing Y Axis Units. The analyzer remembers the settings for both Log and Lin.

AMPLITUDE / Y Scale

AMPLITUDE / Y Scale
:DISPlay:WINDow[1]:TRACe:Y:[SCALe]:SPACing LINear|LOGarithmic
:DISPlay:WINDow[1]:TRACe:Y:[SCALe]:SPACing?
2.0.5 Presel Center
Adjusts the centering of the preselector filter to optimize the amplitude accuracy at the active marker frequency. If no marker is on when Presel Center is pressed, the analyzer turns on the currently selected marker and places it on the peak signal. If a marker is already enabled, it must be placed at the signal peak before activating this function. Presel Center only functions when measuring signals ≥2.85 GHz in band 1 and higher bands. (The instrument is in band 1, or higher, when the stop frequency is >3 GHz.)
For analyzers with a frequency range greater than 26.5 GHz, there are two preselectors (microwave band and millimeter bands). To create an adjustment value for both, the user must setup and request a Presel Center in a microwave band, then set up and request a Presel Center in a millimeter band.
With Input Mixer (Ext) selected and the Mixer Type set to Presel (preselected), Presel Center adjusts the frequency of the external preselector filter to maximize the amplitude at the active marker frequency.
[:SENSe]:POWer[:RF]:PCENter
Remote Command Notes: This command has no effect if it is invoked in a non-preselected band, or the active marker is less than 3 GHz.

AMPLITUDE / Y Scale
Allows you to manually adjust the preselector filter frequency to optimize its response on the signal of interest. This function is only available when measuring signals ≥2.85 GHz in band 1 and higher bands.
For analyzers that have both millimeter wave and microwave types of bands (E4448A and E4446A analyzers) the adjustment is band specific. If the currently active marker, or the center frequency, is in one of the four lower bands (bands 1 to 4) the preselector adjustment specified is used for the microwave bands. However, if the active marker or center frequency is in one of the two upper bands (bands 5 to 6), then the preselector adjustment is applied for those millimeter bands.
When the preselector adjustment is being changed, the active function area of the display shows a third line (mm-wave bands preselector, or MW bands preselector) dependent on the preselector applied. The preselector applied is determined by the analyzer settings of the start, stop, and marker frequency settings as shown below. Note that applied preselector will be notated in the active function block of the display.
PSA Frequency Setting Preselector Applied
[:SENSe]:POWer[:RF]:MW:PADJust
[:SENSe]:POWer[:RF]:MW:PADJust?
[:SENSe]:POWer[:RF]:MMW:PADJust
Microwave · Millimeter

AMPLITUDE / Y Scale
[:SENSe]:POWer[:RF]:MMW:PADJust?
2.0.7 Y Axis Units
Displays the menu keys that enable you to change the vertical (Y) axis amplitude units. The analyzer retains the entered Y Axis Units separately for both Log and Lin amplitude scale types. For example, if Scale Type has been set to Log, and you set Y Axis Units to dBm, pressing Scale Type (Log) sets the Y Axis Units to dBm. If Scale Type has been set to Lin and you set Y Axis Units to Volts, pressing Scale Type (Lin) sets the Y Axis Units to Volts. Pressing Scale Type (Log) again sets the Y Axis units back to dBm.
This key is unavailable (greyed out) when the FM Demod View is on.
Y Axis Units, in conjunction with the Scale Type, affect how the data is read off the display, markers, and over the remote interface. When using the remote interface no units are returned, so you must know what the Y-Axis units are to interpret the results:
For Scale Type (Lin) = Volts
:UNIT:POWer DBM|DBMV|DBUV|V|W
:UNIT:POWer?
Remote Command Notes: The settings of Y Axis Units and Scale Type, affect how the data is read over the remote interface. When using the remote interface no units are returned, so you must know what the Y-Axis units are to interpret the results:
Example 1, set the following:
Scale Type (Log) Y Axis Units, dBm Scale/Div, 1 dB Ref Level, 10 dBm
This sets the top line to 10 dBm with each vertical division representing 1 dB. Thus, if a point on trace 1 is on the fifth graticule line from the top, it represents 5 dBm and will read out remotely as 5.
Example 2, set the following:
Scale Type (Lin) Y Axis Units, Volts Ref Level, 100 mV (10 mV/div)

AMPLITUDE / Y Scale
This sets the top line to 100 mV and the bottom line to 0 V, so each vertical division represents 10 mV. Thus, if a point on trace 1 is on the fifth graticule line from the top, it represents 50 mV and will read out remotely as 50.
2.0.7.1 dBm
Sets the amplitude units to dBm.
See “Y Axis Units” on page 56.
2.0.7.2 dBmV
Sets the amplitude units to dBmV.
See “Y Axis Units” on page 56.
2.0.7.3 dBµV
Sets the amplitude units to dBµV.
See “Y Axis Units” on page 56.
2.0.7.4 Volts
Sets the amplitude units to volts.

AMPLITUDE / Y Scale
See “Y Axis Units” on page 56.
2.0.7.5 Watts
Sets the amplitude units to watts.
See “Y Axis Units” on page 56.
Allows you to add an offset value to the displayed reference level. The reference level is the absolute amplitude represented by the top graticule line on the display. Reference-level offsets are entered by using the numeric keypad or programming commands. The knob and step keys are not active.
Offsets are used when gain or loss occurs between a device under test and the analyzer input. Thus, the signal level measured by the analyzer may be thought of as the level at the input of an external amplitude conversion device. Entering an offset does not affect the trace position or attenuation value, just the displayed value readouts such as reference level and marker amplitudes.
The maximum reference level available is dependent on the reference level offset. That is, Ref Level −Ref Level Offset must be in the range −170 to +30 dBm.
For example, the reference level value range can be initially set to values from −170 dBm to 30 dBm with no reference level offset. If the reference level is first set to −20 dBm, then the reference level offset can be set to values of −50 to +150 dB.
If the reference level offset is first set to −30 dB, then the reference level can be set to values of −200 dBm to 0 dBm. In this case, the reference level is “clamped” at 0 dBm because the maximum limit of +30 dBm is reached with a reference level setting of 0 dBm with an offset of −30 dB. If instead, the reference level offset is first set to 30 dB, then the reference level can be set to values of −140 to +60 dBm.
When a reference level offset is entered, the offset value appears on the left side of the display under Offst (as opposed to frequency offsets which appear at the bottom of the display.) To eliminate an offset, press Ref Lvl Offst, 0, and dB.
Key Notes: Only numeric entries are valid, the knob and step keys are not applicable to this function.
Annunciation/

AMPLITUDE / Y Scale
Annotation: The offset is displayed to left of Trace window; third from the top, just below the scale type.
:DISPlay:WINDow[1]:TRACe:Y:[SCALe]:RLEVel:OFFSet (in dB)
:DISPlay:WINDow[1]:TRACe:Y:[SCALe]:RLEVel:OFFSet?
Example: DISP:WIND:TRAC:Y:RLEV:OFFS 12.7 Sets the Ref Level Offset to 12.7 dB. The only valid suffix is dB. If no suffix is sent, dB will be assumed.
(Option 1DS only.) Turns the internal preamp on and off. The preamp functions over a frequency range of 100 kHz to 3 GHz. When the preamp is on, an automatic adjustment compensates for the gain of the preamp so that displayed amplitude readings still accurately reflect the value at the analyzer input connector. The preamp is switched off for frequencies above 3 GHz, and the correction is not applied, even though the PA annotation remains on screen. For signal frequencies below 100 kHz, the preamp is not automatically switched out, but signal amplitude roll-off occurs even in the “DC” setting of the RF Coupling control.
The gain of the preamp is nominally 30 dB (PSA) or 20 dB (ESA).
[:SENSe]:POWer[:RF]:GAIN[:STATe] OFF|ON|0|1
[:SENSe]:POWer[:RF]:GAIN[:STATe]?
2.0.10 Corrections
Accesses the Corrections menu keys that allow you to enable the corrections function and to

AMPLITUDE / Y Scale
select which set of correction factors you wish to modify. These frequency/amplitude corrections will be applied to the displayed data to correct for system losses/gains outside the analyzer. Four different sets of correction data can be stored.
There is no equivalent remote command for this key.

AMPLITUDE / Y Scale
2.0.10.1 Apply Corrections
Pressing Apply Corrections (Yes) turns on the amplitude-correction factors. Corrections will only be applied for the sets of correction factors whose correction state is set to On. To turn a set of correction factors on, use the Correction On Off key in the Antenna, Cable, Other, or User menus.
Annunciation/ Annotation: When Apply Corrections (Yes) is selected, an A will appear in the screen annotation on the left edge of the display, whether or not a correction set has been turned on using the Correction (On) key in the Antenna, Cable, Other, or User menus.
[:SENSe]:CORRection:CSET:ALL[:STATe] OFF|ON|0|1
[:SENSe]:CORRection:CSET:ALL[:STATe]?
2.0.10.2 Antenna, Cable, Other, and User Keys
Pressing Antenna, Cable, Other, or User accesses the Correction menu for that type of correction data. These 4 keys display the status of correction sets. If the key indicates On, then amplitude corrections for this type have been enabled. To perform the corrections, both this key and Apply Corrections must indicate Yes. The status is toggled in the correction set menu located under Correction.
There is no remote command for this key.

AMPLITUDE / Y Scale
2.0.10.2.1 Correction
Turns the amplitude correction function on or off for the specific set of correction data. The corrections state must be set to On for the correction to be applied.
Key Path: AMPLITUDE / Y Scale, More, Corrections, Antenna (Cable, Other, or User)
[:SENSe]:CORRection:CSET[1]|2|3|4[:STATe] OFF|ON|0|1
[:SENSe]:CORRection:CSET[1]|2|3|4[:STATe]?
Remote Command Notes: [:SENSe]:CORRection:CSET:ALL[:STATe] must be set to on for this command to function.
CSET number equivalents to front-panel access definitions are as follows:
CSET or CSET1 is Antenna CSET2 is Cable CSET3 is Other CSET4 is User
2.0.10.2.2 Edit
Accesses menu keys that allow you to create and edit an amplitude-correction factor set. It puts the analyzer into a split-screen mode where the correction data is displayed in a table under the trace data. Pressing ESC while in this menu will exit the menu and remove the table from the screen. New points will be applied only after the editor is closed.
Key Path: AMPLITUDE / Y Scale, More, Corrections, Antenna (Cable, Other, or User)
[:SENSe]:CORRection:CSET[1]|2|3|4:DATA ,{,,} Creates an amplitude-correction factor set
[:SENSe]:CORRection:CSET[1]|2|3|4:DATA:MERGe ,{,,} Adds the points with the specified values to the current amplitude correction data, allowing you to merge correction data. If too much

AMPLITUDE / Y Scale
data is merged, as many points as possible are merged into the existing data and then an error is reported.
[:SENSe]:CORRection:CSET[1]|2|3|4:DATA?

AMPLITUDE / Y Scale
Remote Command Notes: [:SENSe]:CORRection:CSET:ALL[:STATe] must be set to on for this command to function.
CSET number equivalents to front-panel access definitions are as follows:
CSET or CSET1 is Antenna CSET2 is Cable CSET3 is Other CSET4 is User
2.0.10.2.2.1 Point
Allows you to create or edit an amplitude-correction factor data point. Up to 200 points may be defined for each set. Enter the point number to be created or edited by using the numeric keypad, then press Enter, or use the knob or step keys to move to an existing point. After selecting a point, Frequency becomes active.
Key Path: AMPLITUDE / Y Scale, More, Corrections, Antenna (Cable, Other, or User), Edit
State Saved: Not part of instrument state, saved in a corrections file.
See “Edit” on page 62
2.0.10.2.2.2 Frequency
Allows you to enter the frequency value for an amplitude-correction point. Enter the frequency value by using the numeric keypad. Change the frequency value by using the step keys or the knob. After entering a frequency, Amplitude becomes active.
A frequency coordinate must always be specified for amplitude-correction factors. Amplitude-correction data is sorted in the table by frequency. The sorting occurs immediately after you have entered the frequency value via the front-panel.
Key Path: AMPLITUDE / Y Scale, More, Corrections, Antenna (Cable, Other, or User), Edit
State Saved: Not part of instrument state, saved in a corrections file.
See “Edit” on page 62

AMPLITUDE / Y Scale

AMPLITUDE / Y Scale
2.0.10.2.2.3 Amplitude
Allows you to enter the amplitude value for the current amplitude-correction point. After entering an amplitude, the point number automatically increments and Frequency becomes active to allow entry of the frequency of the next point.
Key Path: AMPLITUDE / Y Scale, More, Corrections, Antenna (Cable, Other, or User), Edit
State Saved: Not part of instrument state, saved in a corrections file.
See “Edit” on page 62
2.0.10.2.2.4 Delete Point
Allows you to delete the amplitude-correction data for the currently selected point. The prompt “If you are sure, press key again to delete” will appear on the display. Pressing Delete Point again will delete the point and adjust all of the point numbers as appropriate.
Key Path: AMPLITUDE / Y Scale, More, Corrections, Antenna (Cable, Other, or User), Edit
See “Edit” on page 62
2.0.10.2.3 Delete Corrections
Allows you to clear all data from the selected amplitude-correction set. The prompt If you are sure, press key again to delete will appear on the display. Pressing Delete again will delete the correction set.
Key Path: AMPLITUDE / Y Scale, More, Corrections, Antenna (Cable, Other, or User)
[:SENSe]:CORRection:CSET[1]|2|3|4:DELete
CSET or CSET1 is Antenna CSET2 is Cable CSET3 is Other CSET4 is User

AMPLITUDE / Y Scale
2.0.10.3 Freq Interp
Allows you to determine how trace values are computed between points in a correction table. If the linear mode is selected, a straight line is used between points in the correction table (for a linear frequency scale and for a decibel amplitude scale). If the logarithmic mode is selected, frequency values between points are computed by first taking the logarithm of both table values and the intermediate value, while using decibel amplitude values.
Defaults.
[:SENSe]:CORRection:CSET[1]|2|3|4:X:SPACing LINear|LOGarithmic
2.0.10.4 Delete All Corrections
Allows you to delete all amplitude-correction sets.
Dependencies/ Couplings: If corrections are on, corrections are turned off.
[:SENSe]:CORRection:CSET:ALL:DELete

AMPLITUDE / Y Scale
Compensates for external gain/loss. The function is similar to the Ref Lvl Offset function, however the value is considered, along with the maximum mixer level setting, to determine the attenuation required (10 dB of Attenuation is added for every 10 dB of External Amp Gain). The gain is subtracted from the amplitude readout so that the displayed signal level represents the signal level at the input of the external device.
Gains may only be entered with the numeric keypad or programming commands, not the knob or step keys.
to the factory default by pressing System, Restore Sys Defaults.
[:SENSe]:CORRection:OFFSet[:MAGNitude] (in dB)
[:SENSe]:CORRection:OFFSet[:MAGNitude]?
Sets the Ext Amp Gain to 7.3 dB. The only valid suffix is dB. If no suffix is sent, dB is assumed.
Permits the selection of 2 dB (for PSA), 5 dB (for ESA), or 10 dB (for PSA/ESA) step resolution for input attenuation.
[:SENSe]:POWer[:RF]:ATTenuation:STEP[:INCRement] (in dB)
[:SENSe]:POWer[:RF]:ATTenuation:STEP[:INCRement]?

AMPLITUDE / Y Scale
Sets the Attenuation to 10 dB. The only valid suffix is dB. If no suffix is sent, dB is assumed.
If a value >5 is entered, 10 is used.
If a value ≤5 is entered, 2 is used for PSA, or 5 is used for ESA.
Enables you to set the relationship between the highest signal that can be displayed (the reference level) and the input attenuation. The relationship applies whenever the Attenuation is set to Auto. The relationship is that the attenuation is given by reference level minus the max mixer level. For example, as the reference level changes, the input attenuator changes to ensure that a signal at the reference level does not exceed the Max Mixer Lvl setting.
[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer]
[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer]?

AMPLITUDE / Y Scale

Coupled functions are functions that are linked/dependent on other functions. Pressing Auto Couple displays some of the most important coupled analyzer functions.
An example of a coupled function is Res BW (resolution bandwidth). When Res BW is set to auto, the Res BW is automatically set based on the span setting.
Coupled functions are affected depending on how they are coupled. For example, Video BW is coupled to Res BW, so changing Res BW affects Video BW, but changing Video BW does not affect Res BW. Changing Video BW puts it in manual (instead of auto-coupled). When Video BW is set to Man it is unaffected by Res BW changes.
When a function is in the Man state, a # will appear next to its annotation on the display. If one or more functions are manually set so that the amplitude or frequency measurement becomes uncalibrated, “Meas Uncal” appears on the top right side of the graticule.
Auto-couples all coupled functions. If Auto All is pressed all coupled functions are set to Auto. Setting any auto coupled function to Man (manual), uncouples that function without changing the coupling of other functions.
Coupled functions are functions that are linked. When Auto All is pressed, the analyzer automatically couples all instrument settings that have auto/man choices. In most cases, these auto settings give the most accurate measurements and optimum dynamic range. When a function is coupled, it is in the Auto state. When it is uncoupled it is in the Man state.
Dependencies/ Couplings: When segmented sweep is on, this key is not available (greyed out).
When Auto All is selected:
Auto Couple

:COUPle ALL|NONE
:COUPle?
Remote Command Notes: sets all the functions to the manual (not coupled) mode. There is no front-panel key equivalent to the COUPle NONE command.
ALL puts all the functions into the auto coupled mode.
Selects the FFT vs. Sweep key functions.
[:SENSe]:SWEep:TYPE AUTO|FFT|SWEep changes the sweep type to FFT or swept, or it lets the analyzer automatically choose the type of analysis to use based on current instrument settings.
[:SENSe]:SWEep:TYPE?
This function is automatically activated when Auto All is selected. Selecting Auto: Best Dynamic Range tells the analyzer to choose between swept and FFT analysis, with a primary goal of optimizing dynamic range. If the dynamic range is very close between swept and FFT, then it chooses the faster one.
While Zero Span is selected, this key is greyed out. The status of the FFT & Swept selection is saved when entering zero span and is restored when leaving zero span.
Auto Couple

[:SENSe]:SWEep:TYPE:AUTO:RULes SPEed|DRANge selects the rules to use when SWE:TYPE AUTO is selected. This setting, combined with your current analyzer setup, is used to select either FFT or swept mode.
[:SENSe]:SWEep:TYPE:AUTO:RULes?
Example: SWEep:TYPE AUTO selects the automatic mode.
SWE:TYPE:AUTO:RUL DRAN sets the rules for the auto mode to dynamic range.
Selecting Auto: Best Speed tells the analyzer to choose between FFT or swept analysis based on the fastest analyzer speed. While Zero Span is selected, this key is greyed out. The auto-couple settings are kept in memory and are restored whenever leaving Zero Span.
[:SENSe]:SWEep:TYPE:AUTO:RULes SPEed|DRANge selects the rules to use when SWE:TYPE AUTO is selected. This setting, combined with your current analyzer setup, is used to select either FFT or swept mode.
See “Auto: Best Dynamic Range” on page 72.
Example: SWEep:TYPE AUTO selects the automatic mode.
SWE:TYPE:AUTO:RUL SPE sets the rules for the auto mode to speed
Manually selects swept analysis, so it cannot change automatically to FFT.
While Zero Span is selected, this key is greyed out. The status of the FFT & Swept selection is saved when entering zero span and is restored when leaving zero span.
Use [:SENSe]:SWEep:TYPE AUTO|FFT|SWEep
See “FFT & Sweep” on page 72.
Auto Couple

Manually selects FFT analysis, so it cannot change automatically to swept.
While Zero Span is selected, this key is greyed out. The status of the FFT & Swept selection is saved when entering zero span and is restored when leaving zero span.
The process of making a spectrum measurement with FFTs is inherently a “gated” process, in that the spectrum is computed from a time record of short duration, much like a gate signal in swept-gated analysis.
The duration of the time record is 1.83 divided by the RBW, within a tolerance of about 3% for bandwidths up through 1 MHz. Therefore, unlike swept gated analysis, the duration of the analysis is fixed by the RBW, not by the gate signal. Because FFT analysis is inherently faster than swept analysis, the gated FFT measurements can have better frequency resolution (a narrower RBW) than would swept analysis for a given duration of the signal to be analyzed.
FFT analysis in the PSA usually involves making auto ranged measurements, and the time required to autorange the FFT can be both long and inconsistent. The PSA hardware automatically sets the ADC Ranging to Bypass when any trigger, except Free Run is selected.
The width of a single FFT measurement can be up to 10 MHz, so gated FFT measurements can only be made for spans of 10 MHz or less.
To make a gated FFT measurement, set the analyzer as follows.
Use [:SENSe]:SWEep:TYPE AUTO|FFT|SWEep
See “FFT & Sweep” on page 72.
Auto Couple

2.1.2.5 FFTs/Span
Displays and controls the number of FFT segments used to measure the entire Span. This key is inactive (greyed out) unless Sweep Type has been set to FFT. If Sweep Type is set to Auto and FFTs are selected, FFTs/Span is still greyed out, and the number of FFTs automatically selected is shown. If Sweep Type is set to Manual:FFT, FFTs/Span becomes available. Press FFTs/Span and an integer can be entered. The analyzer will try to use the number entered, but it may need to use more due to hardware or software limitations.
An FFT can only be performed over a limited span or segment (also known as the FFT width). Several FFT widths may need to be combined to measure the entire span. The “FFT Width” is (Span)/(FFTs/Span), and affects the ADC Dither function. (See Auto Couple).
FFT measurements require that the signal level driving the A/D converter in the IF be small enough to avoid overloading, and that the gain that controls that signal level remain fixed during the measurement of an entire FFT segment. This constraint can allow higher dynamic ranges in swept mode in some cases, but increasing FFTs/Span can restore that dynamic range to FFT measurements, at the expense of losing some of the speed advantages of the FFT.
For example, in pulsed-RF measurements such as radar, it is often possible to make high dynamic range measurements with signal levels approaching the compression threshold of the analyzer in swept spans (well over 0 dBm), while resolving the spectral components to levels below the maximum IF drive level (about -8 dBm at the input mixer). But FFT processing experiences overloads at the maximum IF drive level even if the RBW is small enough that no single spectral component exceeds the maximum IF drive level. If the user reduces the width of an FFT using the FFTs/Span function, an analog filter is placed before the ADC that is about 1.3 times as wide as the FFT segment width. This spreads out the pulsed RF in time and reduces the maximum signal level seen by the ADC. Therefore, the input attenuation can be reduced and the dynamic range increased without overloading the ADC.
Further improvement in dynamic range is possible by changing the ADC gain. In swept analysis in PSA, the gain is normally auto ranged such that it can track the signal power as the analyzer sweeps through CW-like signals. Since FFT processing cannot autorange the gain within the measurement of a single FFT segment, the autoranging advantage is lost for single FFT measurements. But if the segments are reduced in width by using more FFTs/Span, then individual FFT segments can use higher gains, improving dynamic range.
Additional information about selecting FFTs/Span can be found in a product note, "PSA Series Swept and FFT Analysis," literature number 5980-3081EN, available online through http://www.agilent.com
Auto Couple

[:SENSe]:SWEep:FFT:SPAN:RATio
[:SENSe]:SWEep:FFT:SPAN:RATio?
Selects the LO (local oscillator) phase noise behavior for various operating conditions. The selected value is displayed below the £(f) indicator on the left side of the screen. It is preceded by # if PhNoise Opt Auto has been selected.
[:SENSe]:FREQuency:SYNThesis 1|2|3
1, selects optimization of phase noise for frequencies offset 50 kHz from the carrier. 3, selects optimization of LO for fast tuning
[:SENSe]:FREQuency:SYNThesis?
[:SENSe]:FREQuency:SYNThesis:AUTO OFF|ON|0|1
[:SENSe]:FREQuency:SYNThesis:AUTO?
FREQ:SYNT 3, selects optimization for fast tuning
2.1.3.1 Auto
Selects the LO phase noise behavior to optimize speed or dynamic range for various instrument operating conditions.
For PSA: the Auto rules choose Fast Tuning whenever the span ≥10.5 MHz or the Res BW >200 kHz. Otherwise, for spans >141.4 kHz, and for Res BWs >9.1 kHz, the Auto rules choose Optimize £(f) for f >50 kHz. All remaining cases choose Optimize £(f) for f <50 kHz.
See “PhNoise Opt” on page 76.
Auto Couple

2.1.3.2 Optimize £(f) for frequencies < 50 kHz
The LO phase noise is optimized for offsets less than 50 kHz from the carrier, at the expense of phase noise beyond 50 kHz offset.
See “PhNoise Opt” on page 76.
2.1.3.3 Optimize £(f) for frequencies > 50 kHz
Optimizes phase noise for offsets above 50 kHz from the carrier, especially those from 70 kHz to 300 kHz. Closer offsets are compromised and the throughput of measurements (especially remote measurements where the center frequency is changing rapidly), is reduced. There is a graph of typical phase noise performance in the Specifications Guide.
See “PhNoise Opt” on page 76.
2.1.3.4 Optimize LO for Fast Tuning
In this mode, the LO behavior compromises phase noise at all offsets from the carrier below approximately 2 MHz. This allows rapid measurement throughput when changing the center frequency or span.
See “PhNoise Opt” on page 76.
See Detector in Det/Demod menu.
Auto Couple

See Avg/VBW Type in the BW/Avg menu.
Auto Couple

Access the menu to set ADC Dither to On, Off, or Auto.
[:SENSe]:ADC:DITHer[:STATe] OFF|ON|AUTO
[:SENSe]:ADC:DITHer[:STATe]?
2.1.6.1 Auto
Turns the ADC dither to automatic. It then chooses on or off according to which is most likely to be the best selection, based on the other analyzer settings such as span, resolution BW and sweep type (FFT or swept).

Auto Couple

Use [:SENSe]:ADC:DITHer[:STATe] AUTO
2.1.6.2 On
When ADC Dither is On, the linearity of low-level signals is improved. However, the ADC dynamic range is reduced to make room for the dither. As a result, the noise floor of the analyzer is somewhat compromised. So making measurements with ADC dither on gives you better amplitude linearity, but turning ADC dither off gives you a lower noise floor (better sensitivity).
[:SENSe]:ADC:DITHer[:STATe] ON
2.1.6.3 Off
When ADC Dither is Off, the instrument noise floor is improved. If ADC dither is on the ADC dynamic range is reduced to make room for the dither. As a result, the noise floor of the analyzer is somewhat compromised. So making measurements with ADC dither on gives you improved amplitude linearity. but turning ADC dither off gives you a lower noise floor (better sensitivity).
Use [:SENSe]:ADC:DITHer[:STATe] OFF
Accesses the keys to set the ADC ranging to provide for the best signal to noise ratio (Bypass), or for the best FFT speed (Autorange).
Auto Couple

[:SENSe]:ADC:RANGe AUTO|NONE
[:SENSe]:ADC:RANGe?
Example:
2.1.7.1 Autorange
Turns the ADC ranging to automatic which provides the best signal to noise ratio. Auto Couple All sets the ADC ranging to Autorange. Autorange is usually preferred over Bypass.
2.1.7.2 Bypass
Turning the ADC ranging to bypass mode allows higher throughput for FFT measurements. (It can improve FFT measurement speed by 10% to 50%.) Bypass also provides better phase noise, though the effect is negligible when using swept analysis or for offsets below 300 kHz. The £(f) improvement is a few dB in FFT mode at 800 kHz offset.
Bypass allows triggered FFT measurements to occur at the trigger time instead of following an autoranging time. Therefore, whenever the trigger selected is anything other than Free Run, and the Manual:FFT sweep selection is made, the ADC hardware is set to Bypass regardless of the setting of ADC Ranging. See “Making Gated FFT Measurements With Your PSA” on page 74. Another advantage of Bypass is lower IF harmonic distortion, which may be visible with very high signal levels at the input mixer.
Auto Couple

Auto Couple

Activates the resolution bandwidth function, and displays the menu keys that control both the bandwidth and averaging functions.
Enables you to select the 3.01 dB resolution bandwidth (RBW) of the analyzer in 10% steps from 1 Hz to 3 MHz, plus bandwidths of 4, 5, 6, or 8 MHz.
If an unavailable bandwidth is entered with the numeric keypad, the closest available bandwidth is selected.
Sweep time is coupled to RBW. As the RBW changes, the sweep time (if set to Auto) is changed to maintain amplitude calibration.
Video bandwidth (VBW) is coupled to RBW. As the resolution bandwidth changes, the video bandwidth (if set to Auto) changes to maintain the ratio set by VBW/RBW.
When Res BW is set to Auto, resolution bandwidth is autocoupled to span. The ratio of span to RBW is set by Span/RBW (described on page 93). The factory default for this ratio is approximately 106:1 when auto coupled. When Res BW is set to Man, bandwidths are entered by the user, and these bandwidths are used regardless of other analyzer settings.
NOTE In zero span, the auto/manual function of this key is not applicable. When Res BW (Auto) is selected in non-zero span, any changes to Res BW while in zero span will revert to the Auto value when you return to non-zero span. When Res BW (Man) is selected in non-zero span, any changes to Res BW while in zero span will be maintained when you return to non-zero span.
A # mark appears next to Res BW on the bottom of the analyzer display when it is not coupled. To couple the resolution bandwidth, press Res BW (Auto) or Auto All.
NOTE For applications that require 6 dB resolution bandwidths, it is possible to use an equivalent 3 dB resolution bandwidth. Because the analyzer has Gaussian RBW, the equivalent 6 dB bandwidth of any RBW filter can be determined using the following formula: 6 dB RBW = 3 dB RBW x 1.414. For example, if a 6 dB RBW of 100 kHz is required, the equivalent 3 dB RBW Filter would be 100 kHz/1.414 = 70.7 kHz. The closest RBW filter for the analyzer that would be used is 68 kHz.
BW/Avg

[:SENSe]:BANDwidth|BWIDth[:RESolution]
[:SENSe]:BANDwidth|BWIDth[:RESolution]?
[:SENSe]:BANDwidth|BWIDth[:RESolution]:AUTO OFF|ON|0|1
[:SENSe]:BANDwidth|BWIDth[:RESolution]:AUTO?
BWID:AUTO On
2.2.2 Video BW
Enables you to change the analyzer post-detection filter from 1 Hz to 8 MHz in approximately 10% steps. In addition, a wide-open video filter bandwidth (VBW) may be chosen by selecting 50 MHz. Video BW (Auto) selects automatic coupling of the Video BW filter to the resolution bandwidth filter using the VBW/RBW ratio set by the VBW/RBW key.
NOTE Sweep Time is coupled to Video Bandwidth (VBW). As the VBW is changed, the sweep time (when set to Auto) is changed to maintain amplitude calibration. This occurs because of common hardware between the two circuits, even though the Video BW filter is not actually “in-circuit” when the detector is set to Average. Because the purpose of the average detector and the VBW filter are the same, either can be used to reduce the variance of the result.
Although the VBW filter is not “in-circuit” when using the average detector, the Video BW key can have an effect on (Auto) sweep time, and is not disabled. In this case, reducing the VBW setting increases the sweep time, which increases the averaging time, producing a lower-variance trace.
When using the average detector with either Sweep Time set to Man, or in zero span, the VBW setting has no effect and is disabled (greyed out).
A “#” mark appears next to VBW on the bottom of the analyzer display when it is not coupled. To couple the video bandwidth, press Video BW (Auto) (or press Auto All).
BW/Avg

BW/Avg

[:SENSe]:BANDwidth|BWIDth:VIDeo
[:SENSe]:BANDwidth|BWIDth:VIDeo?
[:SENSe]:BANDwidth|BWIDth:VIDeo:AUTO OFF|ON|0|1
[:SENSe]:BANDwidth|BWIDth:VIDeo:AUTO?
BWID:VID:AUTO ON
BWID:VID:AUTO?
2.2.3 VBW/RBW
Selects the ratio between the video and resolution bandwidths. Video bandwidth wider than resolution bandwidth (VBW/RBW ratio > 1.000), provides the best peak measurements of signals such as wideband radar pulses. VBW narrower than RBW (VBW/RBW ratio < 1.000) reduces the variance of noise-like signals and makes spectral components close to the noise floor easier to view. The knob and step keys change the ratio in a 1, 3, 10 sequence. With Preset Type set to Factory, pressing Preset or selecting Auto Couple, Auto All sets the ratio to 1.000 X. When VBW/RBW (Auto) is selected, the ratio is determined as indicated below.
Table 2-1 VBW/RBW Ratio Auto Rules
Detector Mode Noise Marker
Peak On 10 Noise response is estimated for wide VBW case
Negative Peak On 10 Don’t care
Average On 0.1 Narrow VBW for low-sigma marker readout
Normal On 0.1 Don’t care (noise marker makes is not valid with normal detection.)
Sample On 0.1 Narrow VBW for low-sigma marker readout
Peak Off 10 Wide VBW for good impulse BW (pulsed RF)
Negative peak Off 10 Don’t care
Average Off 0.1 Narrow VBW for low-sigma trace results
Normal Off 1.0 Swept analysis CW signal setting
Sample Off 1.0 Wide VBW for good impulse BW (pulsed RF)
BW/Avg · Ratio Notes

[:SENSe]:BANDwidth|BWIDth:VIDeo:RATio
[:SENSe]:BANDwidth|BWIDth:VIDeo:RATio?
[:SENSe]:BANDwidth|BWIDth:VIDeo:RATio:AUTO OFF|ON|0|1
[:SENSe]:BANDwidth|BWIDth:VIDeo:RATio:AUTO?
BAND:VID:RAT:AUTO 0
BAND:VID:RAT:AUTO?
2.2.4 Average
Initiates a digital averaging routine that averages the trace points in a number of successive sweeps, resulting in trace “smoothing.” You can select the number of sweeps (average number) with the numeric keypad (not the knob or step keys). Increasing the average number further smooths the trace. To select the type of averaging used, press BW/Avg, Avg/VBW Type.
Averaging restarts when any of the following occurs:
In single sweep, the specified number of averages is taken, then the sweep stops. In continuous sweep, the specified number of averages is taken, then the averaging continues,
1 Average Number
with each new sweep averaged in with a weight of and the old average
Average Number 1 – Average Number
reduced by multiplying it by .
To turn off averaging, press Average (Off).
BW/Avg

BW/Avg

Remote Commands:
[:SENSe]:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:AVERage[:STATe]?
[:SENSe]:AVERage:COUNt
[:SENSe]:AVERage:COUNt?
[:SENSe]:AVERage:CLEar
Remote Command Notes: For valid average data, you must re-start the trace at the beginning of a sweep. To do this remotely, first abort (:ABORT) the sweep and then initiate a single sweep (:INIT:CONT OFF).
AVER:CLE clears the current average and restarts the averaging process.
Displays the functions that enable you to automatically or manually choose one of the following averaging scales: log-power (video), power (RMS), or voltage averaging.
NOTE When you select log-power averaging, the measurement results are the average of the signal level in logarithmic units (decibels). When you select power average (RMS), all measured results are converted into power units before averaging and filtering operations, and converted back to decibels for displaying. Remember: there can be significant differences between the average of the log of power and the log of the average power.
The following are the averaging processes within a spectrum analyzer, all of which are affected by this setting:
When manual is selected, the type is shown on the left side of the display with a #. When auto is selected, the analyzer chooses the type of averaging. When one of the average types is selected manually, the analyzer uses that type regardless of other analyzer settings, and sets Avg/VBW Type to Man.
BW/Avg

[:SENSe]:AVERage:TYPE RMS|LOG|SCALar
[:SENSe]:AVERage:TYPE?
[:SENSe]:AVERage:TYPE:AUTO OFF|ON|0|1
[:SENSe]:AVERage:TYPE:AUTO?
AVER:TYPE:SCAL Sets Voltage averaging
AVER:TYPE:LOG Sets Log-Power (video) averaging
2.2.5.1 Auto
Chooses the optimum type of averaging for the current instrument measurement settings.
Auto selects Power (RMS) Averaging if Marker Noise is on, Band/Intvl Power is on, or Detector is set to Man and Average. It selects Voltage Averaging if Amplitude, Scale Type is set to Lin. For other conditions, Auto selects Log-Power Average.
See “Avg/VBW Type” on page 89
2.2.5.2 Log-Pwr Avg (Video)
Selects the logarithmic (decibel) scale for all filtering and averaging processes. This scale is sometimes call “Video” because it is the most common display and analysis scale for the video signal within a spectrum analyzer. This scale is excellent for finding CW signals near noise, but its response to noise-like signals is 2.506 dB lower than the average power of those noise signals. This is compensated for in the Marker Noise function. When this type of averaging is selected, LgAv appears on the left side of the display.
The equation for trace averaging on the log-pwr scale is shown below, where N is the number of averages accumulated. (In continuous sweep mode, once N has reached the Average Number, N stays at that value, providing a running average.)
New avg N 1 – ( )Oldavg Newdata + N =
Assumes all values in decibel scale.
BW/Avg

or Auto Couple, Avg/VBW Type
See “Avg/VBW Type” on page 89
2.2.5.3 Pwr Avg (RMS)
In this average type, all filtering and averaging processes work on the power (the square of the magnitude) of the signal, instead of its log or envelope voltage. This scale is best for measuring the true time power of complex signals. This scale is sometimes called RMS because the resulting voltage is proportional to the square root of the mean of the square of the voltage. When this type of averaging is selected, PAvg appears on the left side of the display.
In the equation for averaging on this scale (below), N is the number of averages accumulated. (In continuous sweep mode, once N has reached the Average Number, N stays at that value.)
10 + N log × =
New Avg 10 N 1 – ( ) 10 ×
Assumes all values in dB.
or Auto Couple, Avg/VBW Type
See “Avg/VBW Type” on page 89.
2.2.5.4 Voltage Avg
In this Average type, all filtering and averaging processes work on the voltage of the envelope of the signal. This scale is good for observing rise and fall behavior of AM or pulse-modulated signals such as radar and TDMA transmitters, but its response to noise-like signals is 1.049 dB lower than the average power of those noise signals. This is compensated for in the Marker Noise function. When this type of averaging is selected, VAvg appears on the left side of the display.
In the equation for averaging on this scale (below), N is the number of averages accumulated. (In continuous sweep mode, once N has reached the Average Number, N stays at that value.)

20 + N log × =
New Avg 20 N 1 – ( ) 10 ×
Assumes all values in dB.
or Auto Couple, Avg/VBW Type
See “Avg/VBW Type” on page 89.

Selects the ratio between span and resolution bandwidth. A factory preset sets the ratio to 106:1. The ratio can be changed using the front-panel step keys, knob, or numeric keypad.
[:SENSe]:FREQuency:SPAN:BANDwidth[:RESolution]:RATio
[:SENSe]:FREQuency:SPAN:BANDwidth[:RESolution]:RATio?
[:SENSe]:FREQuency:SPAN:BANDwidth[:RESolution]:RATio:AUTO OFF|ON|0|1
[:SENSe]:FREQuency:SPAN:BANDwidth[:RESolution]:RATio:AUTO?
Example: FREQ:SPAN:BAND:RAT 200 sets a ratio of 200:1, and turns off the auto coupling.
FREQ:SPAN:BAND:RAT?
BW/Avg

BW/Avg

Displays the menu keys that control parameters associated with the detector modes.
Select a specific type of detector, or choose Auto to let the instrument select the appropriate detector for a particular measurement.
When discussing detectors, it is important to understand the concept of a trace “bucket.” For every trace point displayed in swept and zero-span analysis, there is a finite time during which the data for that point is collected. The analyzer has the ability to look at all of the data collected during that time and present a single point of trace data based on the detector mode. We call the interval during which the data for that trace point is being collected, the “bucket.” Thus, a trace is more than a series of single points. It is actually a series of trace “buckets.” The data may be sampled many times within each bucket.
In swept analysis, the time interval of data collection for a display bucket also represents a frequency interval. In FFT analysis, the bucket represents just a frequency interval. The detector determines the relationship between the spectrum computed by the FFT and the single data point displayed for the bucket.
When the Detector choice is Auto, the detector selected depends on marker functions, trace functions, and the trace averaging function.
See “Auto Rules For Detector Selection” on page 97 for information on the Auto detector selection.
When you manually selected a detector (instead of selecting Auto), that detector is used regardless of other analyzer settings.
The detector choices are:
Det/Demod

Because they may not find a spectral component's true peak, neither average nor sample detectors measure amplitudes of CW signals as accurately as peak or normal, but they do measure noise without the biases of peak detection.
The detector in use is indicated on the left side of the display. If the detector has been manually selected, a # appears next to it.
To measure the average power (RMS voltage) in each display point, set Detector to Average, and verify that Avg/VBW Type is set to Pwr Avg (RMS).
[:SENSe]:DETector[:FUNCtion] AVERage|NEGative|NORMal|POSitive|SAMPle|RMS
[:SENSe]:DETector[:FUNCtion]?
The query returns a name that corresponds to the detector mode as shown by the following terms:

2.3.1.1 Auto
The system selects normal detection as the default, but if a condition arises where a different type of detection scheme would be better utilized, the system uses the alternate scheme. For example, when in Auto mode, the Marker Noise function uses Average detection because the system determines that the average detector has a better speed/variance trade-off for noise-type signals.
Refer to Figure 2-1, which shows a decision tree of how detection type is determined.
Figure 2-1 Auto Rules For Detector Selection

Det/Demod

[:SENSe]:DETector:AUTO OFF|ON|0|1
[:SENSe]:DETector:AUTO?
2.3.1.2 Normal
Displays the peak-detected level in the interval (bucket) being displayed when the signal is CW-like. If the signal is noise-like (within a bucket the signal both rose and fell), the even bucket shows the peak (maximum) within a two-bucket interval, and the odd bucket shows the negative peak (minimum). Gain is increased to compensate for the effects of faster sweep rates, to keep the displayed value accurate.
When Normal is selected, Norm appears on the left side of the display.
See “Detector” on page 95
2.3.1.3 Average
For each interval (bucket) in the trace, Average detection displays the average of all samples within the interval using one of the following averaging methods:
To explicitly set the averaging method, use the BW/Avg, Avg/VBW Type key. The combination of the average detector and the power method is equivalent to what is sometimes referred to as “RMS detection.” When the method (Avg/VBW Type) is set to Pwr Avg, and Detector is set to Average, the RMS method is selected.
When Average is selected, Avg appears on the left side of the display.
Dependencies/ Couplings: Use of Average affects the VBW setting. See BW/Avg, VBW. When in Average detection, video trigger is not available.
See “Detector” on page 95.
Det/Demod

2.3.1.4 Peak
For each interval (bucket) in the trace, Peak detection displays the highest amplitude within the interval. Peak detection is used for CW measurements and some pulsed-RF measurements. For swept analysis, peak detection basically obtains the maximum video signal between the end of the last bucket and the start of the next one. Gain is increased to compensate for the effects of faster sweep rates, to keep the displayed value accurate. For FFT analysis, the highest spectral amplitude is displayed, even if that peak amplitude falls between samples of the spectrum computed in the FFT process.
When Peak is selected, Peak appears on the left side of the display.
See “Detector” on page 95
2.3.1.5 RMS (Remote Command Only)
Selects the Average Detector. If BW/Avg, Avg/VBW Type is set to Auto (or Pwr Avg) this will yield the RMS voltage (average power) for each trace point. (See 3.4.1.3, Average)
Key Path: There is no key selection for this setting, but you can access it by using Average Detector (see “Average” on page 98).
See “Detector” on page 95.
2.3.1.6 Sample
The sample detector displays the instantaneous level of the signal at the center of the interval (bucket) represented by each trace point.
Sample detection is primarily used to display noise or noise-like signals.
Sample detection is not best for amplitude measurements of CW-like signals for two reasons. First, the peak response to a signal can occur between samples, so unless the Span to RBW ratio is lower than usual, the highest sample can be well below the peak signal amplitude. Second, for the high sweep rates normally used, the peak response of the RBW filters is up to –0.5 dB. This sweeping error is compensated when using the peak and normal detectors by changing the overall gain. But the gain is not changed when in the sample detector, because to do so would cause errors in the response to noise.
Det/Demod

When Sample is selected, Samp appears on the left side of the display.
Det/Demod

See “Detector” on page 95
2.3.1.7 Negative Peak
For each interval (bucket) in the trace, Negative Peak detection displays the lowest sample within the interval. Negative peak detection is similar to peak detection, but selects the minimum video signal.
When Negative Peak is selected, NPk appears on the left side of the display.
See “Detector” on page 95
Det/Demod

Det/Demod

Displays menu keys that enable you to control certain items on the display of the analyzer.
When Full Screen is pressed the measurement window expands horizontally over the entire instrument display. It turns on/off the display of the softkey labels. Pressing any other key that results in a new menu will cancel the full screen function.
State Saved: Not saved in state.
:DISPlay:FSCReen[:STATe] OFF|ON|0|1
:DISPlay:FSCReen[:STATe]?
Activates an adjustable horizontal line that is used as a visual reference line. The line has an amplitude value that corresponds to its vertical position relative to the reference level. The value of the display line appears on the left side of the display below the label Dl. The display line can be adjusted using the step keys, knob, or numeric keypad. The units of Display Line are determined by the Y-Axis Units setting under Amplitude.
State Saved: Saved in instrument state.
Display

:DISPlay:WINDow:TRACe:Y:DLINe
:DISPlay:WINDow:TRACe:Y:DLINe?
:DISPlay:WINDow:TRACe:Y:DLINe:STATe OFF|ON|0|1
:DISPlay:WINDow:TRACe:Y:DLINe:STATe?
2.4.3 Display Line
Activates an adjustable horizontal line that is used as a visual reference line. The line has an amplitude value that corresponds to its vertical position relative to the reference level. The value of the display line appears on the left side of the display below the label Dl. The display line can be adjusted using the step keys, knob, or numeric keypad. The units of Display Line are determined by the Y-Axis Units setting under Amplitude.
State Saved: Saved in instrument state.
:DISPlay:WINDow:TRACe:Y:DLINe
:DISPlay:WINDow:TRACe:Y:DLINe?
:DISPlay:WINDow:TRACe:Y:DLINe:STATe OFF|ON|0|1
:DISPlay:WINDow:TRACe:Y:DLINe:STATe?
Display

Display

Selects the screen position for the Active Function Display. Depending on the type of trace date being viewed, you can move the Active Function Display position for less visual interference with your screen data.
Readback: The selection of the position to display the active function (Top, Center, Bottom) reads back to line 3 of this key.
State Saved: Not saved in instrument state, survives power cycle and preset.
Factory Preset: No effect, persistent variable, survives power cycle and preset.
Factory Default: Center. Pressing System, Restore Sys Defaults will reset to factory defaults.
:DISPlay:AFUNction:POSition BOTtom|CENTer|TOP
:DISPlay:AFUNction:POSition?
2.4.4.1 Top
Displays the active function in the top-left corner of the display.

Readback: Top reads back to line 3 of this key.
State Saved: Not saved in instrument state, survives power cycle and preset.
Factory Preset: No effect, persistent variable, survives power cycle and preset.
Display

:DISPlay:AFUNction:POSition TOP
2.4.4.2 Center
Displays the active function in the center-left side of the display.

Readback: Center reads back to line 3 of this key.
State Saved: Not saved in instrument state, survives power cycle and preset.
Factory Preset: No effect, persistent variable, survives power cycle and preset.
:DISPlay:AFUNction:POSition CENTer
Display

2.4.4.3 Bottom
Displays the active function in the bottom-left corner of the display.

Readback: Bottom reads back to line 3 of this key.
State Saved: Not saved in instrument state, survives power cycle and preset.
Factory Preset: No effect, persistent variable, survives power cycle and preset.
:DISPlay:AFUNction:POSition BOTtom
Limit lines can be defined to compare the data to your defined limits and indicate a pass or fail condition. Limits accesses menus that allow you to create, modify, and change the properties of limit lines. There are two limit lines in the instrument.
State Saved: Not saved in instrument state.
SCPI Status Bits/ OPC Dependencies: No OPC dependencies.
Factory Default: Unaffected by system defaults.
Display

:CALCulate:LLINe[1]|2:DATA , , {,,,}
:CALCulate:LLINe[1]|2:DATA?
Defines the limit line values, and destroys all existing data. Up to 200 points may be defined for each limit using the following parameters.
can be frequency or time values as specified by the following command: :Calculate:LLINe:CONTrol:DOMain. Frequencies are always in Hz. Time is always in seconds. No unit is allowed in this parameter.
Range: -30 Gs to +30 Gs for time limits, -3 kHz to +350 GHz for frequency limits.
amplitude values are always in units of dBm. Up to two amplitude values can be provided for each x-axis value, by repeating in the data list. No unit is allowed in this parameter.
connected values are either "0" or "1." A "1" means this point should be connected to the previously defined point to define the limit line. A "0" means that it is a point of discontinuity and is not connected to the preceding point. The connected value is ignored for the first point.
:CALCulate:LLINe[1]|2:DATA:MERGe , , {,,,}
Adds the points with the specified values to the current limit line, allowing you to merge limit line data. Up to two amplitude values are allowed for each x value. If more than 200 points are entered to be merged, the first 200 points are merged into the existing limit, then an error ‘too many DATA entries’ is reported.
Remote Command Notes: Up to 200 points total may be defined for each limit.
2.4.5.1 Limit 1 or Limit 2
Selects Limit 1 or Limit 2 for modification.
State Saved: Not saved in instrument state.
Display

2.4.5.1.1 Type (Upper Lower)
Allows you to define the limit you are editing as either an upper or lower limit. An Upper limit fails if the trace exceeds the limit. A Lower limit fails if the trace falls below the limit.
Display, Limits, Limit 2
Dependencies/ Couplings: If a margin has already been set for this limit line, and this key is used to change the limit type, then the margin value is reset to 0 dB.
State Saved: Not saved in instrument state. Survives power cycle and preset. The limit-line data is saved in files.
Factory Preset and *RST: Not affected by preset.
Factory Default: Limits are off by default. Upper for Limit 1, then a limit line is created. Lower for Limit 2, then a limit line is created.
:CALCulate:LLINe[1]|2:TYPE UPPer|LOWer
:CALCulate:LLINe[1]|2:TYPE?
Example: :CALC:LLIN2:TYPE LOW sets limit line 2 as a lower limit.
:CALC:LLIN1:TYPE? responds with the limit line 1 limit type.
2.4.5.1.2 Limit Display
Turns limit-line display On or Off. Either Limit or Margin, as well as Test, must be turned on to turn on a limit test.
Display, Limits, Limit 2
State Saved: Not saved in instrument state.
SCPI Status Bits/ OPC Dependencies: No OPC dependencies.
Display

:CALCulate:LLINe[1]|2:DISPlay OFF|ON|0|1
:CALCulate:LLINe[1]|2:DISPlay? queries the current limit line.
Example: :CALC:LLIN2:DISP OFF turns off the display of the limit lines.
:CALC:LLIN1:DISP? tells you whether the limit lines are being displayed.
2.4.5.1.3 Limit Test
Turns the testing of the limit line On or Off. If the trace is at or within the bounds of the set limit or margin, PASS LIMIT# or PASS MARGIN# is displayed in green in the upper-left corner of the measurement area where # is the number of the selected limit line.
Only positive margins are allowed for lower limits and only negative margins are allowed for upper limits. If the trace is out of the limit or margin boundaries, FAIL LIMIT# or FAIL MARGIN# is displayed in red. The results for Limit 2 are displayed below those for Limit 1. Either Limit or Margin must be turned on for Test to be turned on.
Display, Limits, Limit 2
Dependencies/ Couplings: If either of the limits or margins are turned off, the test cannot be turned on. That is, if both Limit and Margin are set to Off, then the test is turned off automatically.
State Saved: Not saved in instrument state.
:CALCulate:LLINe[1]|2:STATe OFF|ON|0|1 to turn limit lines on or off.
:CALCulate:LLINe[1]|2:STATe?
:CALCulate:LLINe[1]|2:FAIL?
Example: :CALC:LLIN:STATE 1 sets limit line 1 test on.
:CALC:LLIN:STATE? responds with the limit line test status.
Display

:CALC:LLIN:FAIL? Queries the status of the limit-line testing. Returns a "0" if the data passes, and returns a "1" if there is a failure. This query value is valid only if Margin or Test is On.
2.4.5.1.4 Margin (On Off)
Turns margin On or Off. Selecting On allows you to set a limit-line offset for the selected limit line. Only positive margins are allowed for lower limits and only negative margins are allowed for upper limits. The margin lines are displayed in a light grey color. If the limit lines are off and margin is on, the trace is checked against the margin, then a pass or fail margin is displayed. Either Limit or Margin, as well as Test, must be turned on to turn on a limit test.
Display, Limits, Limit 2
Dependencies/ Couplings: If neither of the limits or margins are turned on, the test cannot be turned on. That is, if both Limit and Margin are set to Off, then the test is turned off automatically. If a margin has been set for this limit line, and this key is used to change the limit type, then the margin values is reset to 0 dB.
State Saved: Not saved in instrument state. Survives preset and power cycle.
Factory Default: Off (when a limit line is created). Default value is not affected by a preset.
:CALCulate:LLINe[1]|2:MARGin:STATe OFF|ON|0|1 turns on margins on or off. If the margin and limit display are both turned off, limit test is automatically turned off.
:CALCulate:LLINe[1]|2:MARGin:STATe?
Display

Responds with the margin state; 0 = off 1 = on.
:CALCulate:LLINe[1]|2:MARGin
Defines the amount of measurement margin that is added to the designated limit line.
:CALCulate:LLINe[1]|2:MARGin?
Responds with the margin offset value.
Display

2.4.5.1.5 Edit
Pressing Edit accesses menus for editing limit lines and for accessing the limit-line table editor. Navigation through the limit-line table is achieved by using the front-panel arrow and tab keys. Entering data in each field navigates to the next field. New limit segments will only be applied after the editor is closed.
Pressing Return, or any key that is not associated with the editor, will close the limit-line table editor.
Display, Limits, Limit 2
Factory Preset and *RST: Exits the edit mode.
2.4.5.1.5.1 Point
Up to 200 points may be defined for each limit line using Point. Enter the point number to be created or edited using the numeric keypad, then press Enter, or use the front-panel knob, or step keys to move to an existing point. The step-up key takes you to the next point, while the step-down key takes you to the previous point. After selecting a point, Frequency (or Time, depending on x-axis selection) becomes active.
Display, Limits, Limit 2, Edit
State Saved: Not affected by state.
Display

2.4.5.1.5.2 Frequency (or Time)
Note that this key label changes to Time if X Axis Units has been selected. Pressing this key allows you to enter a value for a limit point in frequency (or time). After entering a value, the limit table is sorted to place the value in the correct order. For a new point, Amplitude defaults to 0 dBm and Connected defaults to Yes. Amplitude then becomes active.
Display, Limits, Limit 2, Edit
Dependencies/ Couplings: Limit-line points are selected according to the X-axis units selected, for example if frequency is selected as your X-axis units, then the limit points are frequency values. Also, if both upper and lower limit lines are selected for the trace then both need to be defined using the same X-axis units.
2.4.5.1.5.3 Amplitude
Pressing Amplitude allows you to enter the amplitude value for the current limit point. After entering a value, Connected becomes active. If a front-panel arrow key is pressed without entering a value, the current Amplitude and Connected values of the point are selected. If the up arrow is pressed, the point number automatically increments to allow entry of the amplitude of the next point. If it is a new point, the Frequency may also be entered for the new point.
Display, Limits, Limit 2, Edit
Display

2.4.5.1.5.4 Connected To Previous Pt
A current point may be connected to the previous point by pressing Yes. No limit testing is performed between disconnected points. Pressing this key when the Connected field is selected toggles the connected value of the current point and increments the Point number to allow entry or editing of the Frequency of the next point. If an arrow key is pressed without entering a value, the current Connected value of the point is selected. If an arrow key is pressed, the Point number automatically increments to allow entry of the Connected value of the next point. If it is a new point, the Frequency may be entered for the new point
Display, Limits, Limit 2, Edit
Knob Increment: Toggles to other selection.
2.4.5.1.5.5 Delete Point
Deleting the current limit point can be achieved by pressing Delete Point. You will be prompted with the message If you are sure, press key again to delete. Pressing Delete Point again will delete the limit point.
Display, Limits, Limit 2, Edit
2.4.5.1.6 Delete Limit
Deleting the current limit set can be achieved by pressing Delete. You will be prompted with the message If you are sure, press key again to delete. Pressing Delete again will delete the limit set.
Display, Limits, Limit 2
:CALCulate:LLINe[1]|2:DELete
Display

Display

2.4.5.1.7 Freq Interp
This key is greyed out if Time is the selected X Axis Units. Sets the interpolation to linear or logarithmic for the specified limiting points set, allowing you to determine how limit trace values are computed between points in a limit table. The available interpolation modes are linear and logarithmic. If the linear (Lin) mode is used for both frequency and amplitude, a straight line is used when interpolating between points in a limit table. If frequency interpolation is logarithmic (Log), frequency values between limit points are computed by first taking the logarithm of both the table values and the intermediate value. A linear interpolation is then performed in this logarithmic frequency space. An exactly analogous manipulation is done for logarithmic amplitude interpolation.
For linear amplitude interpolation and linear frequency interpolation, the interpolation is computed as:
yi 1 + yi – fi 1 + fi – f fi – ( ) yi + =
For linear amplitude interpolation and log frequency interpolation, the interpolation is computed as:
yi 1 + yi – log fi 1 + log fi – log f log fi – ( ) yi + =
For log amplitude interpolation and linear frequency interpolation, the interpolation is computed as:
log yi 1 + log yi – fi 1 + fi – f fi – ( ) log yi + =
For log amplitude interpolation and log frequency interpolation, the interpolation is computed as:
log yi 1 + log yi – log fi 1 + log fi – log f log fi – ( ) y log i + =
Display, Limits, Limit 2
State Saved: Persistent; retains settings, even through a power cycle.

:CALCulate:LLINe[1]|2:CONTrol:INTerpolate:TYPE LOGarithmic|LINear
:CALCulate:LLINe[1]|2:CONTrol:INTerpolate:TYPE?
Example: :CALC:LLIN2:CONT:INT:TYPE LIN sets limit line 2 frequency interpolation to linear.
2.4.5.1.8 Amptd Interp
Allows you to determine how limit trace values are computed between points in a limit table. The available interpolation modes are linear (Lin) and logarithmic (Log). If the linear mode is used for both frequency and amplitude, a straight line is used when interpolating between points in a limit table. This function does not work in zero span (when the analyzer is in time domain). Refer to the "Note" in “Freq Interp” on page 118 for more information.
Display, Limits, Limit 2
State Saved: Persistent; retains settings, even through a power cycle.
Factory Default: Log for a new limit.
:CALCulate:LLINe[1]|2:AMPLitude:INTerpolate:TYPE LOGarithmic|LINear
:CALCulate:LLINe[1]|2:AMPLitude:INTerpolate:TYPE?
Example: :CALC:LLIN:AMPLLINT:TYPE LOG sets limit lines 1 amplitude interpolation to LOG.
:CALC:LLIN:AMPLLINT:TYPE? responds with the limit line interpolation type.
Display

2.4.5.2 X Axis Units
Selects how the limit-line segments are defined. Pressing X Axis Units selects whether the limit lines will be entered using frequency (Freq) or sweep time (Time) to define the segments. They can be specified as a table of limit-line segments of amplitude versus frequency, or of amplitude versus time. A time value of zero corresponds to the start of the sweep, which is at the left edge of the graticule. Switching the limit-line definition between Freq and Time will erase both of the current limit lines. The following message will appear on screen.
Changing X axis units will delete all limits. If you are sure, press key again to change units.
Press X Axis Units again to purge both limit lines and to switch between frequency and time.
Factory Default: Frequency for a new limit.
:CALCulate:LLINe:CONTrol:DOMain FREQuency|TIME
:CALCulate:LLINe:CONTrol:DOMain?
Remote Command Notes: For TIME, the limit line segments are placed on the spectrum analyze display with respect to the sweep time setting of the analyzer, with 0 at the left edge of the display.
For FREQuency, segments are placed according to the frequency that is specified for each segment.
Example: :CALC:LLIN:CONT:DOM FREQ sets limit lines 1 and 2 x-axis units to frequency.
:CALC:LLIN:CONT:DOM TIME sets limit lines 1 and 2 x-axis units to time.
:CALC:LLIN:CONT:DOM? responds with limit lines 1 and 2 x-axis unit type.
Display

2.4.5.3 Limits (Fixed Rel)
Specifies whether the current limit lines are fixed or relative.
Pressing Limits to choose fixed (Fixed) or relative (Rel) limit lines. The fixed (Fixed) type uses the current limit lines as a reference with fixed frequency and amplitude values. The relative (Rel) setting results in the current limit-line value to be relative to the displayed center frequency and reference level amplitude values. When limit lines are specified with time, rather than frequency, the Rel setting only affects the amplitude values. The current amplitude values will be relative to the displayed reference level amplitude, but the time values will always start at the left edge of the graticule.
For example, assume you have a frequency limit line. If the limit line is specified as fixed (Fixed) entering a limit line segment with a frequency coordinate of 300 MHz displays the limit line segment at 300 MHz. If the same limit line table is specified as relative (Rel), it is displayed relative to the analyzer center frequency and reference level. If the center frequency is at 1.0 GHz, a relative limit line segment with a frequency coordinate of 300 MHz will display the limit line segment at 1.3 GHz. If the amplitude component of the relative limit line segment is -10 dB, the - 10 dB is added to the reference level value to obtain the amplitude of the given segment (reference level offset included).
A limit line entered as Fixed may be changed to Rel, and one entered as Rel may be changed to Fixed. When changing between fixed and relative limit lines, the frequency and amplitude values in the limit line table change so that the limit line remains in the same position for the current frequency and amplitude settings of the analyzer. If a time and amplitude limit line is used, the amplitude values change but the time values remain the same.
Factory Default: Fixed for a new limit.
:CALCulate:LLINe:CMODe FIXed|RELative
:CALCulate:LLINe:CMODe?
Example: :CALC:LLIN:CMOD FIX sets limit lines 1 and 2 limits to fixed.
Display

2.4.5.4 Delete All Limits
Deletes the selected limit line. Pressing Delete Limits purges the data from the limit-line tables. Pressing Delete Limits after the prompt, If you are sure, press key again to delete, will delete the limits.
Factory Preset and *RST: Not affected by preset.
:CALCulate:LLINe:ALL:DELete
Example: :CALC:LLIN:ALL:DEL deletes all of the data points for limits lines 1 and 2.
Displays menu keys that enable you to change or clear a title on your display.
2.4.6.1 Change Title
Allows you to write a title across the top of the display. Press Change Title to access the Alpha Editor Menus that contain available characters and symbols. You may also use the numeric keypad to enter numbers. Press Enter or Return to complete the entry. Press ESC to cancel the entry and preserve your existing title.
The display title will remain until you press Change Title again, or you recall a trace or state, or a Factory Preset is performed. A title can also be cleared by pressing Title, Clear Title.
Pressing this key cancels any active function.
State Saved: Saved in instrument state.
:DISPlay:ANNotation:TITLe:DATA
:DISPlay:ANNotation:TITLe:DATA?
Display

2.4.6.2 Clear Title
Allows you to clear a title from the front-panel display. Once cleared, the title cannot be retrieved.
There is no equivalent command, but the example below shows how to enter an empty title.
Accesses menu keys which allow you to turn the graticule and annotation on or off.
State Saved: Saved in instrument state.
2.4.7.1 Graticule (On Off)
Pressing Graticule turns the display graticules On or Off.
:DISPlay:WINDow:TRACe:GRATicule:GRID[:STATe] OFF|ON|0|1
:DISPlay:WINDow:TRACe:GRATicule:GRID[:STATe]?
Display

Display

2.4.7.2 Annotation
Turns the screen annotation on or off for all windows, however, menu key annotation will remain on the display. The screen annotation may not be required for prints or during remote operation.
:DISPlay:WINDow:ANNotation[:ALL] OFF|ON|0|1
:DISPlay:WINDow:ANNotation[:ALL]?
Turns the display on/off. If enable is set to off, the display will appear to blank. This can make the measurement run faster since the instrument does not have to update the display after every data acquisition. There is often no need to update the display information when using remote operation.
Key Path: None, no front-panel control.
State Saved: Not saved in instrument state.
:DISPlay:ENABle OFF|ON|0|1
:DISPlay:ENABle?
Display

Display

Displays a menu of functions that enable you to load, save, and manage data on either a floppy disk (A:) or the analyzer’s internal drive (C:); you can recall, save, copy, delete, or rename files of instrument states, trace data, and screen captures. The menu keys display dialog boxes appropriate for the selected function.
Agilent analyzers use different types of mass storage devices:
The MMEMory command syntax term is a specifier having the form: ‘DRIVE:\DIRECTORY\NAME.EXT’, where the following rules apply:
You can save the following types of files:
Trace + State: A file that contains the trace data and a copy of the current analyzer state. The trace and state are stored in an internal data format (TRC), which cannot be loaded into a PC, but can be loaded back into the analyzer. Traces can be loaded individually or as a group. When a Trace + State file is loaded into the analyzer the trace data that was on the screen, when saved, is loaded into the analyzer. This enables you to view the trace as it looked when it was saved. Because the state data is also saved, the analyzer settings, including all the annotation on the screen, is restored as well. To preserve the trace data, the traces contained in the saved files
File · File Types

are placed in View mode (see Trace/View, page 283) so that they are not immediately overwritten by new trace data. This means that you can save traces while making a measurement, and later load them back into the analyzer, where you can print them or transfer them to a computer, in CSV format, for analysis. If you wish to compare two saved traces, place traces in view mode before saving them. This prevents the trace from being rewritten based on a state change from subsequent loads.
File

CSV: A file that contains trace data in comma-separated values format (CSV, standard PC spreadsheet format), to be read into a spreadsheet for analysis. Most spreadsheet programs support CSV format. They cannot be loaded back into the analyzer.
Units: frequency or time (Display, Limits, Properties, X Axis Units). If you are in time X-Axis Units, and you load frequency limits, all current limit line data will be erased and the analyzer will switch to frequency units. The reverse of the this situation also holds true.
Bitmap: A file that contains an exact bit representation of the screen. Stored in Graphics Interchange Format (GIF) format.
Metafile: A file that contains information about the objects on the screen. Stored in Windows Metafile Format (WMF) format, a format that can be read with Microsoft Word and Microsoft Excel, among others.
Reverse Bitmap: Same as Bitmap, but the black display backgrounds are changed to white and the yellow traces are turned to green to preserve printer black ink.
Reverse Metafile: Same as Metafile, but the black display backgrounds are changed to white and the yellow traces are turned to green to preserve printer black ink.
File

Displays directories and files located on the selected drive, depending upon the preferences set under the Type (page 131) and Sort (page 133) keys. Catalog displays menus to navigate the drives and to sort and select the files you wish to view.
The internal analyzer “drive” (I:\) is used for instrument firmware and optional measurement personalities. It is not available for data/file storage.
:MMEMory:CATalog?
Remote Command Notes: Query returns all files in the specified drive\path name.
The return data will be in the format: , {,}
Each indicates the name and size in bytes of one file in the directory list in the form: “, ” for example, a file called “SCREN000.GIF” which is 21286 bytes in size, would list as “SCREN000.GIF,,21286.” Directories are indicated by square brackets, for example “[MYDIR],,.”
All files are listed, without regard to the preferences selected for the file catalog on the analyzer screen.
If you use lowercase characters, they are converted to uppercase in interpreting catalog commands.
File

2.5.1.1 Type
Allows you to select the desired type of instrument-data files to be displayed. Common types of instrument data files include trace data, limit line data, and amplitude correction data. See “File Types” on page 127 for more information The catalog displays all files (if Type is set to All) or files of the currently selected file type. All directories are always displayed.
Type Format Destination Extension
Trace Trace + state Trace 1, 2, 3, or all traces TRC
Comma separated trace values Trace 1, 2, 3, or all traces CSV
Limit Internal data format LIM
Corrections Internal data format ANT, CBL, OTH, and AMP
Comma separated values CSV
Factory Preset: Type survives Factory Preset and *RST, but is set to State at power on.
There is no remote command for this key.
2.5.1.1.1 All
Displays all files located in the selected directory. If selected, it applies to Catalog, Delete, Copy, and Rename.
2.5.1.1.2 State
Displays all state files (STA) in the selected directory. State files contain most instrument settings. If selected, it applies to all File functions.

File

2.5.1.1.3 Trace
Displays all trace files (TRC and CSV) in the selected directory. If selected, it applies to all File functions.
2.5.1.1.4 Limits
Displays all limits files (LIM) in the selected directory.
2.5.1.1.5 Screen
Displays all screen files (GIF and WMF) in the selected directory.
2.5.1.1.6 Corrections
Displays all correction files (ANT, CBL, OTH, and AMP) in the selected directory.
2.5.1.2 Sort
Displays the Sort menu keys that enable you to view your saved files according to a selected file attribute.The selections include, By Date, By Name, By Extension, By Size, and Order. Order (Up) sorts files in ascending order (for example, A,B,C). Order (Down) sorts files in descending order (for example, C,B,A).
The Sort setting applies to all of the File functions, except Save.
State Saved: The Sort order survives Preset, but is not saved in the instrument state.
There is no remote command for this key.
2.5.1.2.1 By Date
Sorts and displays the current file catalog by the date of the files.
2.5.1.2.2 By Name
Sorts and displays the current file catalog in alphabetical order of the name of the files.
File

File

2.5.1.2.3 By Extension
Sorts and displays the current file catalog, in alphabetical order, by the file extension of the file names (for example: .TRC, .STA).
2.5.1.2.4 By Size
Sorts and displays the current file catalog by the size of the files.
2.5.1.2.5 Order
Changes the order of the display of the current file catalog. Up sorts the files in ascending order (A to Z, 1 to 9), while Down sorts in descending order (Z to A, 9 to 1).
2.5.1.3 Dir Up
Moves up one subdirectory level within a directory. If your position is in the top level of the drive already, it moves up to the drive level and the current drive is highlighted (A: or C:).
2.5.1.4 Dir Select
Selects the drive or directory that is highlighted on the display. You can use the up and down arrows to select and highlight the desired drive or directory. If the top entry in the catalog has a “. .” indication, you are in a subdirectory, and this key acts the same as the Dir Up key. When you are at the top directory level, this key moves up to the drive level.
File

Displays menu keys that enable you to save files to the floppy (A:) or internal (C:) drive.
The menus allow you to fill in data-entry fields for file name, type, format, source, and path (directory). Some fields may be blank depending on file type.
The catalog list box is active and can be used for selecting the directory in which to save the file. Saved files that match the current Type and Format are shown. The Sort Order is always Down, By Date.
Saving directly to floppy drive A:\ can be slow. Try saving first to internal drive C:\ and then transfer the file to the floppy drive.
Factory Preset: State is the default file type at power on.
File

2.5.2.1 Save Now
Executes the save function. While the file is being saved the popup message “Saving file” followed by “Reading directory” is displayed. After a successful save, the text message “xxxxxx file saved” (where xxxxxx is the file name) appears in the status line. Once you have used the File, Save, Save Now keys to setup and save a file, the Save hardkey will perform an immediate Save Now of your file in the same format and to the same location.
The analyzer will pick a filename for you based on the table below. The ### in the auto-generated file name represents a three-digit number which the analyzer has chosen to be the lowest number in the current sequence that does not conflict with an existing file name. The number starts at 000 with a new analyzer or after the installation of new firmware and counts up with each attempted Save. After a Restore Sys Defaults, the number will start at a number that may be lower then the lowest number of the file the currently saved files. The number counts up with each attempted Save, but will skip the numbers already in use and not over write existing files. If you want to enter your own file name, refer to Name (page 143) for additional information.
State STATE### .STA
Trace TRACE### .TRC or .CSV
Screen SCREN### .GIF or .WMF
If the Path: field above the directory box is empty when pressing Save Now, the status line will display the error message: Unable to save file, invalid path. In this case, please select a drive.

:MMEMory:STORe:SCReen
:MMEMory:STORe:STATe 1,
:MMEMory:STORe:TRACe ,
This command is not available when you are outside the Spectrum Analysis mode and working in optional measurement modes.
:MMEMory:STORe:LIMit LLINE1|LLINE2,
:MMEMory:STORe:CORRection ANTenna|CABLe|OTHer|USER,
:MMEMory:STORe:RESults
Trace labels are: TRACE1|TRACE2|TRACE3|ALL The file name must have a file extension of .TRC or .CSV. The file extension determines whether a trace is stored (.CSV), or a trace with its state (.TRC), are stored.
The must include the complete path, for example ‘C:\MYTRACE.TRC’. Lowercase characters are interpreted as uppercase.
These commands will fail if the already exists.
For the MMEM:STOR:LIM LLINE2,”C:mylimit.lim” command, there is no short form for parameters LLINE1|LLINE2.
Example: MMEM:STOR:STAT 1,’C:\mystate.sta’ saves the current instrument state to the specified file name. The .sta extension is required.
*SAV saves the current instrument state to a file name REGxxx, where xxx = the register number. The available register numbers are 0 to 127.
MMEM:STOR:SCR ‘C:\myscreen.gif’ The file must have a .gif or .wmf file extension. The specified file extension determines which file format the instrument will use to save the image. Only Bitmap and Metafile are available (not Reverse Bitmap and Reverse Metafile).
MMEM:STOR:TRAC TRACE3,’C:\mytrace.trc’ Saves trace 3 to the trace + state file C:\MYTRACE.TRC
MMEM:STOR:CORR ANT, ‘A:TEST1.AMP’ saves the current antenna correction to the specified file name. The .amp extension is required.
MMEM:STOR:LIM LLINE2,’C:mylimit.lim’ saves the current limit line two data set to the specified file name. The .lim extension is required.
MMEM:STOR:RES ‘A:ACP.CSV’ saves the current ACP measurement results to the specified file name. The .csv extension is required.
File · *SAV

2.5.2.2 Type
Allows you to select the type of data you want to save. The file types available for saving are described below. See “File Types” on page 127 and “Type” on page 131 for more information.
Type Format Source Extension
Trace Trace + state Trace 1, 2, 3, or all traces TRC
Comma separated trace values Trace 1, 2, 3, or all traces CSV
Limit Internal data format LIM
Corrections Internal data format ANT, CBL, OTH, and AMP
Comma separated values CSV
Factory Preset: Type survives Factory Preset and *RST, but is set to State at power on.
2.5.2.3 Format
When Type is set to Trace, Format allows you to choose between Trace + State and CSV formats. For more information on file types, refer to “File Types” on page 127.
When Type is set to Screen, Format allows you to choose between Bitmap, Metafile, Reverse Bitmap, and Reverse Metafile formats. For more information on file types, refer to “File Types” on page 127.
State Saved: Format is not saved in Instrument State.
Trace file format is Trace + State at power on

Screen file format is Bitmap at power on
File

2.5.2.3.1 Trace + State
When the file type is Trace, this key selects the Trace + State, instrument-readable file (TRC) format for your file. For more information on file types, refer to “File Types” on page 127.
When the file type is Trace, this key selects the trace data as comma-separated values (CSV). The CSV format is readable by a spreadsheet on your computer, but the trace cannot be restored to the analyzer display. For more information on file types, refer to “File Types” on page 127.
2.5.2.3.3 Bitmap
When the file type is Screen, this key selects the bitmap Graphics Interchange Format (GIF) file format for your saved data. For more information on file types, refer to “File Types” on page 127.
2.5.2.3.4 Metafile
When the file type is Screen, this key selects the metafile Windows Metafile Format (WMF) file format for your saved data. For more information on file types, refer to “File Types” on page 127.
2.5.2.3.5 Reverse Bitmap
When the file type is Screen, this key selects the inverse bitmap file format (GIF) for your saved data. For more information on file types, refer to “File Types” on page 127.
2.5.2.3.6 Reverse Metafile
When the file type is Screen, this key selects the inverse metafile file format (WMF) for your saved data. For more information on file types, refer to “File Types” on page 127.
File

2.5.2.4 Source
When the file type is set to Trace, this key allows you to save trace 1, 2, 3 or All. Saving trace All saves all traces in a single.TRC file.
When the file type is set to Corrections, Source accesses the Antenna, Cable, Other and User menu keys, which allow you to select the type of correction to be saved.
When the file type is set to Limits, Source accesses the Limit 1 and Limit 2 menu keys. Limit 1 and Limit 2 provide data sets to determine whether a trace has exceeded preset specifications. Limit sets can hold up to 200 points and can only be saved individually.
For any other Save type, Source is disabled (greyed out).
State Saved: Source is not saved in Instrument State.
Factory Preset: Source survives Factory Preset and *RST, but is set to All Traces at power up.
2.5.2.4.1 Trace 1
Selects trace 1 to be saved.
2.5.2.4.2 Trace 2
Selects trace 2 to be saved.
2.5.2.4.3 Trace 3
Selects trace 3 to be saved.
2.5.2.4.4 All Traces
Selects all the traces to be saved.
File

2.5.2.5 Name
Displays the Alpha Editor and enables you to enter a filename. The numeric keypad can also be used while entering file names. Press Enter or Return to complete the name entry.
The file name is entered as part of the directory/path name that is sent with the SCPI command. See “Save Now” on page 137.
2.5.2.6 Dir Up
Moves up one subdirectory level within a directory. If your position is in the top level of the drive already, it moves up to the drive level and the current drive is highlighted (A: or C:).
The directory is entered as part of the directory/path name that is sent with the SCPI command. See “Save Now” on page 137.
2.5.2.7 Dir Select
Displays the highlighted directory. See “Dir Select” on page 135 for more information.
The directory is entered as part of the directory/path name that is sent with the SCPI command. See “Save Now” on page 137.
Displays the menu key that enables you to load instrument-data files from the selected drive and directory back into the instrument. This function displays the file list box, which shows the data-entry fields for the file name, type, destination, and path.
The catalog list box is active and can be used for selecting the file information in the data-entry fields. Only loadable files that match the current type are shown. Placing the
File

cursor on a file name causes it to be loaded into the file name field.
2.5.3.1 Load Now
Loads the currently selected file. Displayed settings include name, type, destination, and path. While the file is being loaded a popup message is displayed “Loading file.” After a successful load, the text message “xxxxxx file loaded” (where xxxxxx is the file name) appears in the status line. When traces are loaded they always load in View mode.
Traces save in TRC format can be loaded individually or as a group. When a trace is loaded, the state that existed when that trace was saved is loaded along with the trace. Also, the loaded trace(s) is/are placed in view mode.
:MMEMory:LOAD:STATe 1, loads the specified state file into the current active state of the instrument.
*RCL loads the state from the specified internal register into the current active state of the instrument. The available register numbers are 0 to 127.
:MMEMory:LOAD:TRACe ,
:MMEMory:LOAD:CORRection ANTenna|CABLe|OTHer|USER,
:MMEMory:LOAD:LIMit LLINE1|LLINE2,
For the MMEM:LOAD:STAT 1 command:
File

Example: MMEM:LOAD:STAT 1,’C:MYSTATE.STA’ loads the state file C:\MYSTATE.STA.
MMEM:LOAD:TRAC TRACE3,’C:MYTRACE.TRC’ loads the trace in file C:\MYTRACE.TRC into trace 3.
File

2.5.3.2 Type
Enables you to select the type of file you want to load. See “File Types” on page 127 and “Type” on page 131 for more information.
The file types available for loading are described in the following table:
Type Format Destination Extension
Trace Trace + state Trace 1, 2, 3, or all traces TRC
Comma separated trace values Trace 1, 2, 3, or all traces CSV
Limit Internal data format LIM
Corrections Internal data format ANT, CBL, OTH, and AMP
2.5.3.3 Sort
Allows you to view saved files according to a selected file attribute. See “Sort” on page 133 for more information.
2.5.3.4 Destination
When Type is set to Trace, Destination allows you to direct your data to Trace 1, Trace 2, or Trace 3 for a single-trace file. If the data is for all three traces (Source was All when they were saved), the data will be returned to the original trace registers, regardless of the Destination setting.
When Type is set to Limits, Destination allows you to direct your data to Limit 1 or Limit 2.
State Saved: Not saved in Instrument State.
Factory Preset: Trace file format, is All Traces at power on.
2.5.3.4.1 Trace 1
Selects trace 1 for the trace data to be loaded into.
File · State State STA

State Saved: Not saved in Instrument State.
Factory Preset: Not affected by Preset. Power up and Restore Sys Defaults sets Trace 1.
2.5.3.4.2 Trace 2
Selects trace 2 for the trace data to be loaded into.
State Saved: Not saved in Instrument State.
Factory Preset: Not affected by Preset. Power up and Restore Sys Defaults sets Trace 1.
2.5.3.4.3 Trace 3
Selects trace 3 for the trace data to be loaded into.
State Saved: Not saved in Instrument State.
Factory Preset: Not affected by Preset. Power up and Restore Sys Defaults sets Trace 1.
2.5.3.5 Dir Up
Moves up one subdirectory level within a directory. If your position is in the top level of the drive already, it moves up to the drive level and the current drive is highlighted (A: or C:).
State Saved: Not saved in Instrument State.
Factory Preset: Trace file format, is All Traces at power on.
2.5.3.6 Dir Select
Displays the highlighted directory. See “Dir Select” on page 135 for more information.
State Saved: Not saved in Instrument State.
Factory Preset: Trace file format, is All Traces at power on.
Displays the Delete menu keys that enable you to delete instrument data files from the selected directory. The catalog list box is active and can be used for selecting file information for the data-entry fields. Only files that match the current type are shown.
File

Placing the cursor on a file name causes it to be loaded into the file name field.
File

2.5.4.1 Delete Now
Executes the delete function. After you select the file or directory you want to delete, press Delete Now to perform the delete. While the file is being deleted, the popup message “Deleting file” followed by “Reading directory” are displayed. After a successful deletion, the text message “xxxxxx file deleted” (where xxxxxx is the file name) appears in the status line.
If you select a directory or subdirectory to delete, the following popup message is displayed “WARNING: You are about to delete all of the contents of directory xxxxxx. Press Delete Now again to proceed or any other key to abort.” (xxxxxx is the full path and directory name).
To quickly delete all of the file in a directory, select the file at the top of the list and press Delete Now repeatedly until all the files are deleted.
:MMEMory:DELete deletes a file.
:MMEMory:RDIRectory deletes a directory.
Remote Command Notes: If does not exist, a “File Name Error” occurs.
and must include the complete path. Lowercase characters are read as uppercase.
Example: MMEM:DEL ‘C:\destinat.trc’ removes the file C:\DESTINAT.TRC.
MMEM:RDIR ‘C:\myDir’ removes directory C:\MYDIR and all files and subdirectories within that directory.
2.5.4.2 Type
Allows you to select the type of file you want to delete. See “File Types” on page 127 and “Type” on page 131 for more information.
Allows you to select the type of files to be displayed for you to delete. Common types of instrument data files include trace data, limit line data, and amplitude correction data. The catalog displays all files, if Type is set to All or files of the currently selected file type. All directories are always displayed.
2.5.4.3 Sort
Allows you to view your saved files according to a selected file attribute. See “Sort” on page 133 for more information.
File

File

2.5.4.4 Dir Up
Moves up one subdirectory level within a directory. If your position is in the top level of the drive already, it moves up to the drive level and the current drive is highlighted (A: or C:).
2.5.4.5 Dir Select
Displays the highlighted directory. See “Dir Select” on page 135 for more information.
Displays the functions to copy instrument data files in the selected directory to the directory and file name that you choose. This key also displays a catalog of the files that are currently saved in the selected directory and data-entry fields for the following: file name, type, and path location.
2.5.5.1 Copy Now
Executes the copy function, coping data files from one directory to another on one or more mass storage devices, using the currently displayed file settings. While the file is being copied, the “Copying file” followed by “Reading directory” popup message is displayed. After a successful copy, the green text message “xxxxxx file copied” (where xxxxxx is the file name) appears in the status line. If a copy is being done for a file that already exists in the “To” directory, the text message “File already exists” appears in the status line.
:MMEMory:COPY ,
Remote Command Notes: The file names must include the complete file paths. Lowercase characters are read as uppercase.
The original file is , and the new copy of the file is .
Example: :MMEM:COPY ‘C:\oldname.sta’,’A:\newname.sta’ copies C:\OLDNAME.STA to A:\NEWNAME.STA.
File

2.5.5.2 Type
Enables you to select the type of file you want to copy. See “File Types” on page 127 and “Type” on page 131 for more information. If Type is set to All, the catalog displays all files, otherwise the files of the currently selected file type are displayed. All directories are always displayed.
Trace Trace + state TRC
Comma separated trace values CSV
Limit Internal data format LIM
Corrections Internal data format ANT, CBL, OTH, and AMP
Comma separated values CSV
2.5.5.3 Sort
Allows you to view your saved files according to a selected file attribute. See “Sort” on page 133 for more information.
2.5.5.4 Dir From/To
Allows you to select the source and destination directories for your copy on one or more drives. When you press Dir From/To, it toggles between the two displayed directory list windows. Allowing you to define the “From” and “To” locations for copying.
State Saved: Powers up with C:\ as both the “From” and “To” drives. Not save in state. Survives Factory Preset.

2.5.5.5 Dir Up
Moves up one subdirectory level within a directory. If your position is in the top level of the drive already, it moves up to the drive level and the current drive is highlighted (A: or C:).
2.5.5.6 Dir Select
Displays the highlighted directory. See “Dir Select” on page 135 for more information.
Allows you to rename a file. The catalog list box is active and can be used for selecting both the path and a file name. Only loadable files that match the current type are shown. Placing the cursor on a file name causes it to be loaded into the file name field.
2.5.6.1 Rename Now
Executes the rename function. When the rename is complete, the message XXXXXX file renamed to YYYYYY (where XXXXXX and YYYYYY are the filenames) will appear in the status line on your display. If you try to rename a file with a name that already exists, the text message (File already exists) appears in the status line. Placing the cursor on a file name causes it to be loaded into the file name field.
:MMEMory:MOVE ,
Remote Command Notes: must include the complete path, and the case must match that of the file to be renamed. must contain the complete path of the destination, and the case of any directories in the path must match those of the directories in the destination path. The case of the destination file name is always interpreted as uppercase.
You can use this command to move files between directories and drives, even though there is no way to do this from the front panel.
File

2.5.6.2 Type
Enables you to select the type of file you want to rename. See “File Types” on page 127 and “Type” on page 131 for more information.
If Type is set to All, the catalog displays all files, otherwise the files of the currently selected file type are displayed. All directories are always displayed.
Trace Trace + state TRC
Comma separated trace values CSV
Limit Internal data format LIM
Corrections Internal data format ANT, CBL, OTH, and AMP
Comma separated values CSV
2.5.6.3 Sort
Allows you to view your saved files according to a selected file attribute. See “Sort” on page 133 for more information.
2.5.6.4 Name
Displays the Alpha Editor and enables you to enter the file name you want to rename the file to. The numeric keypad can also be used to enter a filename while the alpha editor is accessed. Complete your entry by pressing Return or Enter. See “Name” on page 143 for more information.

2.5.6.5 Dir Up
Moves up one subdirectory level within a directory. If your position is in the top level of the drive already, it moves up to the drive level and the current drive is highlighted (A: or C:).
2.5.6.6 Dir Select
Displays the highlighted directory. See “Dir Select” on page 135 for more information.
Displays the functions to create a new subdirectory in the currently selected directory.
2.5.7.1 Create Dir Now
Executes the create a new directory function. While the directory is being created a popup message is displayed “Creating directory” followed by “Reading directory.” After the successful creation of a directory, the text message “Directory xxxxxx created” (where xxxxxx is the new directory name) appears in the status line. If the creation of a new directory is being performed for a directory name that already exists, the text message “Directory already exists” appears in the status line.
:MMEMory:MDIRectory
Remote Command Notes: must contain the complete path for the new directory. Lowercase characters are interpreted as uppercase.
Example: MMEM:MDIR ‘C:\myDir’ creates directory MYDIR on the C:\ drive.
2.5.7.2 Name
Displays the Alpha Editor and enables you to enter a directory name. The numeric keypad can also be used to enter a directory name while the alpha editor is accessed. To complete the entry, press Return or Enter.
File

File

2.5.7.3 Dir Up
Moves up one subdirectory level within a directory. If your position is in the top level of the drive already, it moves up to the drive level and the current drive is highlighted (A: or C:).
2.5.7.4 Dir Select
Displays the highlighted directory. See “Dir Select” on page 135 for more information.
Deletes all the files on a floppy disk; any information on the disk will be destroyed.
Key Notes: This key displays the file manager display form which includes data entry fields for the new drive name and path.
There is another definition of Format key that defines the format (file type) of you data that you want to save.
There is no remote command for this key.
2.5.8.1 Delete All Now
Executes the Delete All function. After pressing Delete All, the following message will appear on the display: WARNING: You are about to destroy ALL data on volume A: Press Delete All again to proceed or any other key to abort. While deleting, a popup message is displayed “Deleting All.” After a successful floppy disk file deletion, the green text message “Volume A: delete complete”, appears in the status line.
There is no remote command for this key.
File

This command queries trace data from the specified trace. The data format is set by the command :FORMat [:TRACe][:DATA]. When ASCII format is selected, the data is comma-separated ASCII values. Real or Integer format uses a definite length block of data.
The number of trace points returned is set by [:SENSE]:SWE:POIN (from 101 to 8192).
:TRACe[:DATA]? TRACE1|TRACE2|TRACE3|LLINE1|LLINE2
Remote Command Notes: Commands :MMEM:STOR:TRAC and :MMEM:LOAD:TRAC are used to transfer trace data to, or from, the internal hard drive or floppy drive of the instrument.
The query returns the current values of the designated trace. The data is terminated with (for GPIB that is newline, or linefeed, followed by EOI set true; for RS-232 this is newline only.)
This command loads a block of data in the format into the instrument memory location . The query form of the command returns the contents of the file identified by , in the format of a definite length block of data. The query can be used for copying files out of the analyzer over the remote bus.
A definite length block of data starts with an ASCII header that begins with # and indicates how many additional data points are following in the block. Suppose the header is #512320.
:MMEMory:DATA ,
:MMEMory:DATA?
Example: MMEM:DATA ‘C:\DEST.TXT’,’#14abcd’ Loads the data “abcd” into C:\DEST.TXT.
MMEM:DATA? ‘C:\SCREN001.GIF’ Initiates a transfer of data from file C:\SCREN001.GIF.
File

File

This command selects the binary data byte order for data transfer. It controls whether binary data is transferred in normal or swapped mode. Normal mode is when the byte sequence begins with the most significant byte (MSB) first, and ends with the least significant byte (LSB) last in the sequence: 1|2|3|4. Swapped mode is when the byte sequence begins with the LSB first, and ends with the MSB last in the sequence: 4|3|2|1.
State Saved: Survives Preset but not power cycle. Not saved in Instrument State.
:FORMat:BORDer NORMal|SWAPped
:FORMat:BORDer?
This command changes the format of the data. It specifies the format used for trace data during data transfer across any remote port. REAL and ASCII formats will format trace data in the current amplitude units. The format of state data cannot be changed. It is always in a machine readable format only.
When in Spectrum Analysis mode using older instrument firmware, you were only allowed to change the format of trace type data that was returned using TRACe[:DATA]. With this old firmware, other types of measurement data was only available in the ASCII format. (That is, data returned using FETCh, MEASure and READ commands while in Signal Analysis mode.)
For corrected trace data (:TRACe[:DATA] with parameter ), REAL and ASCII formats will provide trace data in the current amplitude units. INTeger format will provide trace data in mdBm. The fastest mode is INTeger,32. However, some measurement data will not fit in 32-bit integers.
ASCII - Amplitude values are in amplitude units separated by commas. ASCII format requires more memory than the binary formats. Handling large amounts of this type of data, takes more time and storage space.
Integer,32 - Binary 32-bit integer values in internal units (dBm), in a definite length block.
Real,32 (or 64) - Binary 32-bit (or 64-bit) real values in amplitude units, in a finite length block. Transfers of real data are done in a binary block format.
File

A definite length block of data starts with an ASCII header that begins with # and indicates how many additional data points are following in the block. Suppose the header is #512320:
State Saved: Saved in Instrument State.
Factory Preset: Survives Preset but not power cycle. Powers up in ASCII format.
:FORMat[:TRACe][:DATA] ASCii|INTeger,32|REAL,32|REAL,64|UINTeger,16
:FORMat[:TRACe][:DATA]?
Remote Command Notes:
Corrected Trace Data Types for :TRACe:DATA?<trace_name>
UINTeger,32 (fastest) Internal Units

File

Displays the menu of frequency functions. Depending on the Frequency entry mode, either the center frequency or the start and stop frequency values appear below the graticule on the display. In Center/Span mode, the Center Frequency and Span appear, and the Center Freq function is automatically activated. In Start/Stop mode, the Start and Stop frequencies appear, and the Start Freq function is automatically activated.
Activates the function that sets the center of the displayed frequency range to the specified frequency.
Annunciation/ Annotation: Center appears in the lower left corner of the display.
Dependencies/ Couplings: Center frequency and span are coupled to each other.
State Saved: Saved in instrument state.
Factory Preset:
E4440A 13.255 GHz
E4443A 3.355 GHz
E4445A 6.605 GHz
E4446A 22.005 GHz
E4448A 25.005 GHz
FREQUENCY / Channel · Model Center Frequency

Range:
Model Frequency Range (with Frequency Offset = 0 Hz)
E4440A –100.0 MHz to 26.0 GHz
E4443A –100.0 MHz to 7.2 GHz
E4445A –100.0 MHz to 13.7 GHz
E4446A –100.0 MHz to 44.5 GHz
E4448A –100.0 MHz to 51.0 GHz
[:SENSe]:FREQuency:CENTer |UP|DOWN
[:SENSe]:FREQuency:CENTer?
FREQ:CENT UP changes the center frequency to 5.1 GHz if you use FREQ:CENT:STEP 100 MHz to set the center frequency step size to 100 MHz
2.6.2 Start Freq
Sets the frequency at the left side of the graticule and sets the frequency entry mode to Start/Stop. When the Start/Stop Frequency entry mode is activate, the start and stop frequency values are displayed below the graticule in place of center frequency and span. The left and right sides of the graticule correspond to the start and stop frequencies.
When Start Freq reaches the upper frequency limit, the stop frequency is set to the highest available frequency and the start frequency is changed to be less then the stop frequency by the minimum span (10 Hz). Center Freq will be updated to the stop frequency minus one-half of the span, or to the stop frequency minus 5 Hz, and Res BW and VBW will be set to 1 Hz.
Dependencies/ Couplings: Start Freq is unavailable when Segmented sweep is on.
FREQUENCY / Channel

Range:
E4440A –100.0000000 MHz to 26.99999999 GHz
E4443A –100.0000000 MHz to 7.19999999 GHz
E4445A –100.0000000 MHz to 13.69999999 GHz
E4446A –100.0000000 MHz to 44.49999999 GHz
E4448A –100.0000000 MHz to 50.99999999 GHz
[:SENSe]:FREQuency:STARt
[:SENSe]:FREQuency:STARt?
2.6.3 Stop Freq
Sets the frequency at the right side of the graticule and sets the frequency entry mode to Start/Stop. When the Start/Stop Frequency entry mode is activate, the start and stop frequency values are displayed below the graticule in place of center frequency and span. The left and right sides of the graticule correspond to the start and stop frequencies.
Factory Preset:
E4440A 26.50000000 GHz
E4443A 6.70000000 GHz
E4445A 13.20000000 GHz
E4446A 44.00000000 GHz
E4448A 50.00000000 GHz
FREQUENCY / Channel · Model Frequency Range · Model Stop Frequency

Range:
E4440A –100.0000000 MHz to 27.0 GHz
E4443A –100.0000000 MHz to 13.7 GHz
E4445A –100.0000000 MHz to 7.2 GHz
E4446A –100.0000000 MHz to 44.5 GHz
E4448A –100.0000000 MHz to 51.0 GHz
[:SENSe]:FREQuency:STOP
[:SENSe]:FREQuency:STOP?
2.6.4 CF Step
Changes the step size for the center frequency and start/stop frequency functions. Once a step size has been selected and the center frequency function is activated, the step keys (and the UP|DOWN parameters for Center Frequency from remote commands) change center frequency by the step-size value. The step size function is useful for finding harmonics and sidebands beyond the current frequency span of the analyzer. When auto-coupled in a non-zero span, the center frequency step size is set to 10% of the span.
Knob Increment:
if Span = 0 Hz increment = RBW/20
if Span > 0 Hz increment = Span/200
Step Key Increment:
if CF Step = auto, Span = 0 Hz step = RBW
if CF Step = auto, Span > 0 Hz step = Span/10
if CF Step = manual step = 1, 2, 5 sequence
FREQUENCY / Channel · Model Frequency Range

Range:
E4440A 1 Hz 27.0 GHz
E4443A 1 Hz 7.2 GHz
E4445A 1 Hz 13.7 GHz
E4446A 1 Hz 44.5 GHz
E4448A 1 Hz 51.0 GHz
[:SENSe]:FREQuency:CENTer:STEP[:INCRement]
[:SENSe]:FREQuency:CENTer:STEP[:INCRement]?
[:SENSe]:FREQuency:CENTer:STEP:AUTO OFF|ON|0|1
[:SENSe]:FREQuency:CENTer:STEP:AUTO?
FREQ:CENT:STEP 500 MHz
FREQ:CENT UP increases the current center frequency value by 500 MHz
FREQ:CENT:STEP:AUTO?
2.6.5 Freq Offset
Enables you to input a frequency offset value to account for frequency conversions external to the analyzer. This value is added to the display readout of the marker frequency, center frequency, start frequency, stop frequency and all other absolute frequency settings in the analyzer. When a frequency offset is entered, the value appears below the center of the graticule. Offsets may only be entered using the numeric keypad, not the knob or step keys. To eliminate an offset, perform a Factory Preset or set the frequency offset to 0 Hz.
This command does not affect any bandwidths or the settings of relative frequency parameters such as delta markers or span. It does not affect the current hardware settings of the analyzer, but only the displayed frequency values. Offsets are not added to the frequency count readouts. Entering an offset does not affect the trace display.
Dependencies/ Couplings: Frequency offset is not available when segmented sweep is on or when frequency scale type is set to log (i.e. Log Sweep is on).
FREQUENCY / Channel · Model Minimum Maximum

[:SENSe]:FREQuency:OFFSet
[:SENSe]:FREQuency:OFFSet?
When a marker is placed on a signal and Signal Track is pressed, the marker will remain on the signal while the analyzer retunes the center frequency to the marker frequency. The analyzer will keep the signal at the center of the display, as long as the amplitude of the signal does not change by more than 3 dB from one sweep to another. If no marker is active, pressing Signal Track to On will activate a marker, perform a peak search, and center the marker on the display.
If the signal is lost, an attempt will be made to find it again and continue tracking. If there are other signals on screen near the same amplitude, one of them may be found instead. Signals near 0 Hz cannot be tracked effectively as they cannot be distinguished from the LO feedthrough, which is excluded by intent from the search algorithm.
When Signal Track is On and the span is reduced, an automatic zoom is performed and the span is reduced in steps so that the signal remains at the center of the display. If the span is zero, signal track cannot be activated.
Signal tracking can also be used with the ∆(delta) pair marker function. This could be used to measure the amplitude at a frequency offset, relative to the fundamental signal amplitude, even if the fundamental is drifting. In this situation, the “tracking” function will be done on the delta marker, not on the reference marker. So you would want to put the ref marker on the frequency+offset, and put the ∆marker on the fundamental signal peak. Then turn on signal tracking. The frequency difference between the two markers will stay fixed. The ∆marker, on the fundamental, will track to the center of the display. The marker readout value will show the amplitude delta between the two markers.
Annunciation/ Annotation: ST appears near the lower-left corner of the display. An (*) may appear in the upper-right corner of the display while the analyzer whenever the analyzer finds that it must retune in order to center the signal on the
FREQUENCY / Channel

display.
FREQUENCY / Channel

Signal Track is not available when any of the following is true:
The following features are not available when Signal Track is on:
State Saved: Saved in instrument state.
:CALCulate:MARKer[1]|2|3|4:TRCKing[:STATe] OFF|ON|0|1
:CALCulate:MARKer[1]|2|3|4:TRCKing[:STATe]?
Example: CALC:MARK1:TRCK ON turns on Signal Track using Marker 1.
CALC:MARK1:TRCK?.
FREQUENCY / Channel

Displays the keys that control some of the signal inputs and outputs of the analyzer.
Brings up a menu of input signal sources, the most common one being the front panel RF Input port.
[:SENSe]:FEED RF|AREFerence|EMIXer
[:SENSe]:FEED?
Example: FEED AREF selects the 50 MHz amplitude reference as the signal input.
Selects the front panel RF Input port to be the analyzer signal input.
See “Input Port” on page 171
Example: FEED AREF selects the 50 MHz amplitude reference as the signal input.
2.7.1.2 Amptd Ref
Selects the 50 MHz, –25 dBm internal amplitude reference as the input signal.
See “Input Port” on page 171
Example: FEED AREF selects the 50 MHz amplitude reference as the signal input.
Input/Output

Specifies alternating current (AC) or direct current (DC) coupling at the analyzer RF input port. Selecting AC coupling switches in a blocking capacitor that blocks any DC voltage present at the analyzer input. This decreases the input frequency range of the analyzer, but prevents damage to the input circuitry of the analyzer if there is a DC voltage present at the RF input.
In AC coupling mode, signals less than 20 MHz are not calibrated. You must switch to DC coupling to see calibrated frequencies of less than 20 MHz. Note that the message DC Coupled will be displayed on the analyzer when DC is selected.
Some amplitude specifications apply only when coupling is set to DC. Refer to the appropriate amplitude specifications and characteristics for your analyzer.
Dependencies/ Couplings: Not available on 40 GHz or 50 GHz analyzers.
:INPut:COUPling AC|DC
:INPut:COUPling?
Sets the input impedance for voltage-to-power conversions. The impedance you select is for computational purposes only, since the actual impedance is set by internal hardware to 50 ohm. The default is 50 ohm. Setting the computational input impedance to 75 ohm is useful when using a 75 ohm to 50 ohm adapter to measure a 75 ohm device on an analyzer having a 50 ohm input impedance.
Input/Output

[:SENCe]:CORRection:IMPedance[:INPut][:MAGNitude] 50|75
[:SENCe]:CORRection:IMPedance[:INPut][:MAGNitude]?
Example: CORR:IMP 75 sets the input impedance correction to 75 ohms.
2.7.4 Trig Input
Sets the external trigger input source to the front panel trigger input connector (EXT TRIGGER INPUT Ext 1) or the rear panel trigger input connector (TRIGGER IN Ext2).
Annunciation/ Annotation: At the left side of the display outside the graticule, the letter “E” indicates that the External Trigger is active.
See “Trig” on page 287
Example: TRIG:SOUR EXT2 to select rear panel external triggering.
Input/Output

Input/Output

This chapter provides key descriptions and programming information for the front-panel key functions of your analyzer starting with the letters M through O. The front-panel functions are listed alphabetically and are described with their associated menu keys. The lower-level menu keys are arranged and described as they appear in your analyzer.


Accesses the marker control menu. If no markers are active, pressing Marker activates the currently selected marker as a normal type marker and places it at the center of the display. (This will be marker 1 if you have not previously selected a different marker.) There are five control modes for the markers:
Normal (POSition) - A single marker that can be moved to any point on the trace.
Delta (DELTa) - A fixed reference marker and a moveable marker that you can place at any point on the trace
Delta Pair (BAND) - Both a movable delta and a movable reference marker. You can independently adjust the position of each marker.
Span Pair (SPAN) - A moveable reference and a movable delta marker. You can adjust the center point of the markers and the frequency span between the markers.
Off (Off) - Turns off the active marker or marker pair.
Your instrument stores data to a high degree of resolution and accuracy. It is often difficult to read the trace data directly from the screen to the desired accuracy. Markers are diamond-shaped pointers that can be placed at any point on a trace to accurately read the data at that point. Markers may also be use in pairs to read the difference (or delta) between two data points. The marker number is indicated above the marker. Use the data controls to position the marker. The knob and/or Up/Down keys move the marker left or right. If a value is entered from the numeric keypad, the marker is moved to the trace point nearest to that value.
The data for the active marker (the one currently be controlled) appears in the upper-right corner of the display. In addition, when a marker is being actively controlled, the marker data appears in the active function area of the display. There are four markers in your instrument; each can be controlled as a single marker or as a reference/delta pair.
A trace is a connected series of points displayed on the instrument screen. The left-most point is point 0 and the right-most point (default) is 600. You control markers by moving them from trace point to trace point. Markers are shaped like diamonds. The lowest point of the diamond shape represents the trace point that is being read. The marker number is indicated above the active marker. The same marker number is indicated with an R (for example, 1R) above the reference marker when in a delta mode (delta, delta pair, and span pair).
Marker · Marker Units

volts. For example, when the percent ratio with Y-axis units in volts is 20% (0.2), the percent ratio with Y-axis units in watts will be 4% (0.22 = 0.04). When you read the value out remotely you have to know whether you are in log (dB) or linear (percent).
Marker

:CALCulate:MARKer[1]|2|3|4:MODE POSition|DELTa|BAND|SPAN|OFF
:CALCulate:MARKer[1]|2|3|4:MODE?
Sets or queries the marker control mode (see parameter list above).
:CALCulate:MARKer[1]|2|3|4:X
Sets the marker X position to a specified point on the X axis in the current X-axis units (frequency or time). If the frequency or time chosen would place the marker off screen, the marker will be placed at the left or right side of the display, on the trace. This command will have no effect if the marker is OFF.
:CALCulate:MARKer[1]|2|3|4:X?
Queries the marker X position in the current x-axis units. The marker must be ON for the response to be valid.
[:SENSe]:MARKer[1]|2|3|4:X:POSition
Sets the marker X position to a specified point on the X axis in display points (values of 0 to 600, or the current number of points in the sweep). The marker must already be on.
[:SENSe]:MARKer[1]|2|3|4:X:POSition?
Returns the current marker X position in display points.
:CALCulate:MARKer[1]|2|3|4:Y?
Queries the marker Y value or delta in the current y axis units. Can also be used to read the results of marker functions such as Marker Noise. The marker must be ON for the response to be valid.
Remote Command Notes: The :CALC:MARK:PEAK:SEARC:MODE MAX|PAR command specifies how a peak is identified for use with the marker commands. See “Peak Search” on page 218.
Example: CALC:MARK:MODE POS selects marker 1 and sets it to Normal.
CALC:MARK2:X 20 GHZ selects marker 2 and moves it to 20 GHz. (Marker 2 must first be turned on.)
Marker

Selects one of the four possible marker or marker pairs. Once a marker is selected, it can be set to any of the control modes, Normal, Delta, Delta Pair, Span Pair, or Off.
State Saved: The number of the selected marker is saved in instrument state.
:CALCulate:MARKer[1]|2|3|4:STATe OFF|ON|0|1
:CALCulate:MARKer[1]|2|3|4:STATe?
Sets or queries the state of a marker. Setting a marker to state ON or 1 selects that marker. Setting a marker which is OFF to state ON or 1 puts it in Normal mode and places it at the center of the display. Setting a marker to state OFF or 0 selects that marker and turns it off. The response to the query will be 0 if OFF, 1 if ON.
Example: CALC:MARK2:STAT ON selects marker 2.
CALC:MARK:STAT ON will not modify a marker that is already on.
Sets the control mode for the selected marker to Normal (see “Marker” on page 177). If the marker is off, a single marker is activated at the center of the display. The marker stays on the trace at the horizontal screen position where it was placed unless Signal Track, or a “marker to” key function (such as Mkr →CF, Mkr →RL, Mkr →CF STEP, Mkr ∆Span, or Min Search) is selected. If you are in a marker pair mode, for example Delta Marker, the reference marker is turned off. You can then adjust the trace point of the marker.
See “Marker” on page 177 for the mode command.
Example: :CALC:MARK:MODE POS selects marker 1 and sets it to Normal.
Marker

Sets the control mode for the selected marker to Delta (see “Marker” on page 177). In Delta mode the display shows the difference between the active (Delta) marker and a reference marker. When Delta mode is selected the reference marker is placed at the current marker position. If the marker is OFF both the active marker and the reference marker are placed at the center of the display. The amplitude and frequency (or time) of the first marker is fixed. The marker number is indicated above the delta marker, and the same number is indicated with an R (for example, 1R) above the reference marker. You can adjust the trace point of the active delta marker. Annotation in the active function block and in the upper-right corner of the display indicates the frequency or time difference and amplitude difference of the two markers. If marker noise is set to On while using Delta and the noise marker is placed on the noise floor, the marker readout displays signal to noise.
Selecting Delta while already in Delta mode causes the reference marker to be reset to the current active (∆) marker position, enabling you to make delta measurements from differing reference points without having to turn off the markers and begin again. Pressing Marker Normal moves the Reference Marker to the Delta Marker position and turns off Delta Marker.
The amplitude of the reference marker is fixed. In non-zero spans the frequency of the reference marker is fixed. If the center frequency of the analyzer is changed such that the reference marker is off the screen, an arrow will appear with the marker number at the left or the right side of the display. This indicates where the trace point is for the reference marker.
In Zero Span the reference marker remains fixed at the trace point on which it was placed. Also, changing Center Frequency does not move the reference marker while in Zero Span.
The markers will be turned off if the scale type is changed between log and linear.
See “Marker” on page 177 for the mode command.
Example: CALC:MARK4:MODE DELT selects marker 4 as a delta marker and places a reference marker at the marker 4 position. If marker 4 is OFF it places both the active and the reference markers at the center of the display.
Sets the control mode for the selected marker to Delta Pair (see “Marker” on page 177). In Delta Pair mode the display shows the difference between the delta marker and a reference marker and enables you to adjust both the Ref (start) and Delta (stop) markers independently. After you turn on the delta Pair function, pressing the key again toggles between the two markers you are controlling. When Ref is underlined you are controlling the reference marker. When ∆is underlined you are controlling the delta marker. The start marker number is indicated with a number and an R above the marker (for example, 1R) and the delta marker is indicated with a marker number.
Marker

There are four conditions that can occur when Delta Pair mode is selected.
The difference between Delta Pair and Delta modes is that in Delta Pair mode the reference marker stays on the trace and you can adjust its trace point. The note (Tracking Ref) appears on the Delta Pair key because, in effect, the reference marker “tracks” the trace. (By comparison, in Delta mode the reference marker does not track changes in the trace results, it remains anchored in amplitude and frequency.)
Once positioned, the markers stay on the trace points you have selected. Ref and Delta markers maintain their displayed x-axis location, but not their frequency values when you change a parameter that redefines the x-axis scale. Reset these markers when parameters such as Span or Center Freq are changed. Adjusting the Span changes the difference between the two markers. Changing the Center changes the center point of the two markers. This function is useful in functions such as Band Power. Changing the frequency or sweep time of the analyzer does not change the trace point of the markers. You cannot move the markers off the screen.
Factory Preset: Ref is the active parameter. Factory preset marker mode is Off.
Range: Refer to the [:SENSe]:SWEep:POINts command under “Points” on page 244.
See “Marker” on page 177 for the command to select the control mode.
:CALCulate:MARKer[1]|2|3|4:X:POSition:STARt
:CALCulate:MARKer[1]|2|3|4:X:POSition:STOP
Sets the reference (Start), or delta (Stop) marker X position to a specified point on the X axis in display points (values of 0 to 600, or the current number of points in the sweep). The marker must already be on.
:CALCulate:MARKer[1]|2|3|4:X:POSition:STARt?
:CALCulate:MARKer[1]|2|3|4:X:POSition:STOP?
Returns the reference (Start), or delta (Stop) marker X position in display points.
Example: CALC:MARK3:MODE BAND activates or selects marker 3 and sets it to Delta Pair. Refer to “Marker” on page 177.
CALC:MARK3:X:POS:STAR 0 moves the reference marker 3 to the left edge
Marker

of the display.
Marker

Sets the control mode for the selected marker to Span Pair (see “Marker” on page 177). In Span Pair mode the display shows the difference between the delta marker and a reference marker and enables you to adjust both the ref and delta markers. The start marker number is indicated with a number and an R above the marker (for example, 1R) and the stop marker is indicated with a marker number. After you turn on the Span Pair function, pressing the key again toggles between the two marker parameters you are controlling, span and center.
Adjusting the Span (Span is underlined) changes the frequency difference between the two markers. Adjusting Center (Center is underlined) maintains the marker spacing and changes the frequency of the midpoint between the markers. Adjusting the span changes the frequency difference between the two markers while maintaining the midpoint between the two markers at a fixed frequency. Changing the center changes the center point between the two markers while maintaining the frequency difference.
There are four conditions that can occur when Span Pair mode is selected.
The difference between Span Pair and Delta modes is that in Span Pair mode the reference marker stays on the trace and you can adjust its trace point.
Once positioned, the markers stay on the trace points on which they have been placed. Changing the frequency or time of the analyzer does not change the trace point of the markers, that is, they stay at the same horizontal position on the display.
You cannot move the markers off the screen. If you adjust either center or span to a value that would cause one of the markers to move off screen, the marker will be placed at the right or left side of the display, on the trace. Changing the Center marker changes the center point of the two markers. This function is useful in functions such as Band Power.
Range: Refer to the [:SENSe]:SWEep:POINts command under “Points” on page 244.
See “Marker” on page 177 for the command to select the control mode.
:CALCulate:MARKer[1]|2|3|4:X:POSition:CENTer
Sets the mid point of the markers to a specific trace point.
:CALCulate:MARKer[1]|2|3|4:X:POSition:CENTer?
Marker

Returns the midpoint trace point.
:CALCulate:MARKer[1]|2|3|4:X:POSition:SPAN
Sets the spacing between the markers to a specified number of trace points.
:CALCulate:MARKer[1]|2|3|4:X:POSition:SPAN?
Returns the spacing of the markers in trace points.
:CALCulate:MARKer[1]|2|3|4:X:CENTer
Sets the mid point of the markers to a specific frequency with a range that matches the units of the trace on which the markers are positioned.
:CALCulate:MARKer[1]|2|3|4:X:CENTer?
Returns the midpoint frequency.
:CALCulate:MARKer[1]|2|3|4:X:SPAN
Sets the spacing between the markers to a specified frequency with a range that matches the units of the trace on which the markers are positioned.
:CALCulate:MARKer[1]|2|3|4:X:SPAN?
Returns the spacing of the markers in frequency.
Example: CALC:MARK3:MODE SPAN selects marker 3 and sets it to Span Pair.
CALC:MARK4:X:POS:SPAN 200 sets the spacing between the markers to 200 trace points for marker pair 4.
CALC:MARK2:X:POS:CENT 300 sets the midpoint between the markers to the 300th trace point from the left of the display. For a 601 point trace this will be the middle of the display.
Turns off the selected marker. In addition, Off turns off functions related to the selected marker such as Signal Track, Band/Intvl Power, and Marker Noise and removes marker annotation from the display.
See “Select Marker” on page 180 for the command to select the control mode.
Example: CALC:MARK3:STAT OFF selects marker 3 and sets it to Off.
Selects the trace that you want your marker or marker pair to be placed on. You can pick
Marker

Marker Trace 1, 2, or 3, or Auto. In Auto mode, the analyzer places markers on the lowest-numbered trace that is in Clear Write mode. If no trace is in Clear Write mode, it places the marker on the lowest-numbered trace in Max Hold mode. If there are none, then it uses the lowest-number trace in Min Hold mode, then in View mode. For example, if trace 1 is in view, and trace 2 is in clear write, any new marker is assigned to trace 2.
State Saved: The Marker Trace for each marker is saved in instrument state.
:CALCulate:MARKer[1]|2|3|4:TRACe 1|2|3
Puts the marker on the specified trace and turns Auto OFF for that marker.
:CALCulate:MARKer[1]|2|3|4:TRACe?
The query returns the number of the trace on which the marker currently resides, even if that marker is in Auto mode.
:CALCulate:MARKer[1]|2|3|4:TRACe:AUTO OFF|ON|0|1
Turning Auto off sets the Marker Trace value to the number of the trace on which the marker currently resides.
:CALCulate:MARKer[1]|2|3|4:TRACe:AUTO?
The response to the query will be 0 if OFF, 1 if ON.
Example: CALC:MARK1:TRAC 2 places marker 1 on trace 2.
This access a menu that enables you to affect how the x-axis information for the selected marker is displayed in the marker area (top-right of display) and the active function area of the display. It only affects the readout on the display of the horizontal position information (for example, frequency).
State Saved: In instrument state, for each marker.
Factory Preset: Frequency for non-zero spans and Time for zero spans.
:CALCulate:MARKer[1]|2|3|4:X:READout FREQuency|TIME|ITIMe|PERiod
Marker

:CALCulate:MARKer[1]|2|3|4:X:READout?
Example: CALC:MARK3:X:READ TIME sets the marker 3 Readout to Time.
3.1.8.1 Frequency
Sets the marker readout to Frequency, displaying the absolute frequency of a normal marker or the frequency of the delta marker relative to the reference marker. Frequency readout is the default setting in non-zero spans and is not available in zero spans.
See “Readout” on page 186 for this command.
Example: CALC:MARK2:X:READ FREQ sets the marker 2 Readout to Frequency.
3.1.8.2 Period
Sets the marker readout to Period, displaying the reciprocal of the frequency at the marker position, or the reciprocal of the frequency separation of the two markers in a delta-marker mode. Period readout is not available in zero spans. If the markers are at the same frequency in a delta marker mode, the result will be the reciprocal of 0, which is infinitely large. The display will show a very large number.
See “Readout” on page 186 for this command.
3.1.8.3 Time
Sets the marker readout to Time, displaying the time interval between a normal marker and the start of a sweep or the time of the delta marker relative to the reference marker. Time is the default setting in zero spans. With a span of zero, the time value is the time position relative to the start of the sweep. In a delta-marker mode it is the (sweep) time interval between the two markers.
See “Readout” on page 186 for this command.
Marker

3.1.8.4 Inverse Time
Sets the marker readout to Inverse Time, displaying the reciprocal of (sweep) time between two markers. This function is only available when in both zero span and in a delta-marker modes. If the markers are at the same x position, the time between them is 0, so the reciprocal of sweep time is infinitely large. The display will show a very large number.
See “Readout” on page 186 for this command.
Marker

When set to On the display is split into a measurement window and a marker data display window. For each marker pair, information is displayed in the data display window, which includes the marker number, trace number, marker type, X axis value, and the amplitude of the marker or the delta value, if a delta marker, or the function value, if in a marker function such as Marker Noise or Band/Intvl Power.
:CALCulate:MARKer:TABLe:STATe OFF|ON|0|1
:CALCulate:MARKer:TABLe:STATe? returns 1 if ON or 0 if OFF.
Example: CALC:MARK:TABL:STAT ON turns on the marker table.
Turns off all markers, including markers used for signal track. This key also turns off marker related functions such as Signal Track, Band Interval Power, and Marker Noise.
:CALCulate:MARKer:AOFF
Example: CALC:MARK:AOFF turns off all markers.
Marker

Marker

Access special marker functions such as frequency counting and noise markers.
:CALCulate:MARKer[1]|2|3|4:FUNCtion BPOWer|NOISe|OFF
:CALCulate:MARKer[1]|2|3|4:FUNCtion?
See “Select Marker” on page 180
Example: CALC:MARK2:STAT ON selects marker 2.
Activates a noise marker for the selected marker. If the selected marker is off it is turned on and located at the center of the display. Reads out the average noise level, normalized to a 1 Hz noise power bandwidth, around the active marker. The noise marker averages 5% of the trace data values, centered on the location of the marker.
The data displayed (if the marker is in Normal mode) is the noise density around the marker. The value readout is followed by “(1 Hz)” to remind you that display is normalized to a one Hz bandwidth.
To measure carrier to noise ratio, be sure that the Marker Fctn is not Marker Noise. Select a Marker, Normal type marker. Place the marker on the signal peak, then select Delta marker. Now place the active (∆) marker on the noise, and select Marker Noise to change the marker type. In this case, the reference marker has units of amplitude and the data displayed is the ratio of the noise density at the delta marker to the reference marker power. The value readout is dB/Hz if the Y-axis units are logarithmic, and % if the Y-axis units are linear. It is understood, in this case, that % stands for the units for volts units and %/Hz for watts units.
To measure the ratio of the noise densities at two locations, be sure that the Marker Fctn is Marker Noise. (The noise is averaged over a region that is 5% of the span, centered at the marker location.) Select Marker, Normal before selecting Delta marker. Then move the active (∆) marker to the second noise location. In this case both markers have units of noise density (for example, dBm/Hz), so the data displayed represents the ratio of the noise density at the delta marker to the noise density at the reference marker. The value readout
Marker Fctn · % Hz ⁄

is displayed as a ratio (dB or %).
Marker Fctn

To guarantee accurate data for noise-like signals, a correction for equivalent noise bandwidth is made by the analyzer. The Marker Noise function accuracy is best when the detector is set to Average or Sample, because neither of these detectors will peak-bias the noise. The trade off between sweep time and variance of the result is best when Avg/VBW Type is set to Power Averaging. Auto coupling, therefore, normally chooses the Average detector and Power Averaging. The Marker Noise function works with all settings of detector and Avg/VBW Type, but using the positive or negative peak detectors does not gives less accurate measurement results.
Note that the value when the Y-axis units are watts is the square of the value when the Y-axis units are volts. For example, when the percent ratio with Y-axis units in volts is 20% (0.2), the percent ratio with Y-axis units in watts will be 4% (0.22 = 0.04). When you read the value out remotely you have to know whether you are in log (dB) or linear (percent), and if linear, whether volts or watts.
Dependencies/ Couplings: Video triggering is not available when the detector is Average, therefore marker functions that would set the detector to Average, and thus conflict with video triggering, are not available when the Video trigger is On.
See “Marker Fctn” on page 191 for the command to select a function.
Example: CALC:MARK:FUNC NOIS turns on marker 1 as a noise marker.
CALC:MARK:FUNC? returns the current setting of marker function for the marker specified. In this case it returns the string: NOIS.
CALC:MARK:Y? returns the y-axis value of the Marker Noise function for marker 1 (if Marker Noise is ON for marker 1).
Measures the power in a bandwidth (non-zero span) or time interval (zero span) specified by the user. If no marker is on, this key activates the delta pair marker mode. If the detector mode is set to Auto, the average detector is selected. If the Avg/VBW type is set to Auto, Power Averaging is selected, other choices of detector and Avg/VBW type will usually cause measurement inaccuracy. The active marker pair indicate the edges of the band. Only Delta Pair and Span Pair marker control modes can be used while in this function, selecting any other mode (for example, Normal or Delta) turns off this function.
Dependencies/ Couplings: Video triggering is not available when the detector is Average, therefore, marker functions that would set the detector to Average, and thus conflict with video triggering, are not available when the Video trigger is On.
Selecting Band/Intvl Power when the marker control function is off, normal, or delta will set the marker control function to delta pair.
Marker Fctn

See “Marker Fctn” on page 191 for the command to select the function.
Example: CALC:MARK:FUNC BPOW turns on marker one as a band power marker.
CALC:MARK:Y? returns the value of the Band/Intvl Power function for marker 1 (if Band/Intvl Power is ON for marker 1).
Turns off marker functions (Band/Intvl Power and Marker Noise).
See “Marker Fctn” on page 191 for the command to select the function.
Example: CALC:MARK2:FUNC OFF turns the marker 2 function off.
Accesses the marker count menu.
3.2.5.1 Marker Count
Turns the marker frequency counter on and off for any active marker. If no marker is active before Marker Count is pressed, a marker is activated at the center of the display. An asterisk (*) may appear in the upper-right area of the display along with the message Cntr 1 (the number in the message depends on the active marker). If the marker count function is on and you change the active marker, the new active marker will use marker count. If the frequency counter function is on with only one active marker and that marker is turned off, then the frequency counter function is turned off. If the frequency counter function is on with multiple markers turned on and only one is turned off, the frequency counter function stays on. Marker Count frequency readings are not affected by the frequency offset function.
In Zero Span the counter continues to function, counting any signal near the center frequency of the analyzer.
Marker Fctn

State Saved: If Marker Count is on, that setting is saved in the instrument state.
:CALCulate:MARKer[1]|2|3|4:FCOunt[:STATe] OFF|ON|0|1
:CALCulate:MARKer[1]|2|3|4:FCOunt[:STATe]?
:CALCulate:MARKer[1]|2|3|4:FCOunt:X?
Marker Fctn

Remote Command Notes: Using the CALC:MARK[1]|2|3|4:FCO command.
If the specified marker number in the command is not the active marker, it becomes the active marker. If the marker number is not turned on, it is first turned on and then it becomes the active marker. Once the marker count function is turned on, it will be on for any active marker, not just the marker number specified when the command was sent.
Using the CALC:MARK[1]|2|3|4:FCO:X? query.
The query returns a 1 only if the marker count function is on and the marker number selected is the currently active marker. The query returns 9e15 if the marker count function is off, or if the specified marker is not the active marker.
CALC:MARK2:FCO:X? returns the counted frequency.
3.2.5.2 Gate Time
Controls the length of time during which the frequency counter measures the signal frequency. For 2 ms and longer gate times, the counter resolution is 0.001 Hz. Longer gate times allow for greater averaging of signals whose frequency is “noisy”, at the expense of throughput. If the gate time is an integer multiple of the length of a power-line cycle (20 ms for 50 Hz power, 16.67 ms for 60 Hz power), the counter rejects incidental modulation at the power line rate. The shortest gate time that rejects both 50 and 60 Hz modulation is 100 ms, which is the value chosen when gate time is in Auto.
State Saved: Saved in instrument state.
:CALCulate:MARKer:FCOunt:GATetime:AUTO OFF|ON|0|1
:CALCulate:MARKer:FCOunt:GATetime:AUTO?
:CALCulate:MARKer:FCOunt:GATetime
:CALCulate:MARKer:FCOunt:GATetime?
CALC:MARK:FCO:GAT 1e-2 sets the gate time to 10-2 s = 10 ms.
Marker Fctn

Accesses menu keys that can copy the current marker value into other instrument parameters (for example, Center Frequency).
Sets the center frequency of the analyzer to the frequency of the selected marker. The marker stays at this frequency, so it moves to the center of the display. This function is not available in Zero Span.
:CALCulate:MARKer[1]|2|3|4[:SET]:CENTer
Example: CALC:MARK2:CENT sets the CF of the analyzer to the value of marker 2.
Sets the center frequency (CF) step size of the analyzer to the marker frequency, or in a delta-marker mode, to the frequency difference between the delta and reference markers. The step size can be verified by pressing Frequency, Center Freq. The step size is displayed in the third line of the active function area of the display. This function is not available in Zero Span.
:CALCulate:MARKer[1]|2|3|4[:SET]:STEP
Example: CALC:MARK1:STEP sets the CF step to the value (or delta value) of marker 1.
Changes the start frequency to the frequency of the active marker. The marker stays at this frequency, so it moves to the left of the display. This function is not available in Zero Span.
:CALCulate:MARKer[1]|2|3|4[:SET]:STARt
Marker ->

Example: CALC:MARK1:STAR sets the start frequency to the value (or delta value) of marker 1.
Marker ->

Changes the stop frequency to the frequency of the active marker. The marker stays at this frequency, so it moves to the right of the display. This function is not available in Zero Span.
:CALCulate:MARKer[1]|2|3|4[:SET]:STOP
Example: CALC:MARK1:STOP sets the stop frequency to the value (or delta value) of marker 1.
Sets the start and stop frequencies to the values of the delta markers. The marker is then set to normal at the center frequency. Only available in Delta, Span Pair, and Delta Pair modes, this function is not available if the marker is off, or in Normal mode or when the span is Zero Span.
:CALCulate:MARKer[1]|2|3|4[:SET]:DELTa:SPAN
Remote Command Notes: Select the delta marker mode with :CALCulate:MARKer[1]|2|3|4:MODE DELTa.
Example: CALC:MARK2:DELT:SPAN sets the start and stop frequencies to the values of the delta markers of marker 2.
Sets the delta marker to the center frequency. Only available in Delta, Span Pair, and Delta Pair modes, this function is not available if the marker is off, or in Normal mode or when the span is Zero Span.
:CALCulate:MARKer[1]|2|3|4[:SET]:DELTa:CENTer
Remote Command Notes: Select the delta marker mode with :CALCulate:MARKer[1]|2|3|4:MODE DELTa.
Example: CALC:MARK2:DELT:CENT sets the center frequency to the value of the delta marker center frequency of marker 2.
Marker ->

Sets the reference level to the amplitude value of the active marker, moving the marked point to the reference level (top line of the graticule).
:CALCulate:MARKer[1]|2|3|4[:SET]:RLEVel
Example: CALC:MARK2:RLEV sets the reference level of the analyzer to the amplitude of marker 2.
Marker ->

Meas Setup (SA with Measurements Off)
Displays the setup menu for the currently selected measurement. This menu is empty if no measurement is active. This could be because Meas Off is selected in the Measure menu.
There is no equivalent remote command.

Meas Setup (SA with Measurements Off)

3.5 MODE
Selects the measurement mode of your analyzer. Spectrum Analysis mode is for general purpose measurement use. The instrument comes with the Spectrum Analysis mode. Additional measurement modes can be added to your instrument memory. Refer to the individual measurement personality mode manuals for instructions on how to install the software.
Dependencies/ Couplings: Other modes, besides Spectrum Analysis, must be installed/licensed in your instrument before they will appear in the Mode menu. Some modes also require the presence of specific hardware.
Saved State: Saved in instrument state.
status bits should be read, and any errors resolved, prior to switching modes. Error conditions that exist prior to switching modes cannot be detected using the condition registers after the mode change. This is true unless they recur after the mode change, although transitions of these conditions can be detected using the event registers.
Changing modes resets all SCPI status registers and mask registers to their power-on defaults. Hence, any event or condition register masks must be re-established after a mode change. Also note that the power up status bit is set by any mode change, since that is the default state after power up.
If Preset Type Mode is selected, then the analyzer settings are preset but it stays in that selected mode.
:INSTrument[:SELect] BASIC|CDMA|CDMA1XEV|CDMA2K|EDGEGSM|LINK|NADC|NFIGURE|PDC|PNOISE|SA|WCDMA
:INSTrument[:SELect]?

Remote Command Notes:
Select the measurement mode. The actual available choices depend upon which modes (measurement applications) are installed in the instrument.A list of the valid choices is returned with the INST:CAT? query.
Once an instrument mode is selected, only the commands that are valid for that mode can be executed.
BASIC CDMA1XEV (1xEV-DO) CDMA2K (cdma2000) EDGEGSM (GSM with EDGE) LINK (89600 VSA Link software) NADC NFIGURE (noise figure) PDC PNOISE (phase noise) SA TDSCDMA WCDMA (3GPP)
3.5.1 Spectrum Analysis
Selects the spectrum analysis measurement mode for your analyzer.
:INSTrument[:SELect] SA
3.5.2 Application Mode Number Selection (Remote command only)
Select the measurement mode by its mode number. The actual available choices depend upon which applications are installed in your instrument.
Dependencies/ Couplings: Other modes, besides Spectrum Analysis, must be installed/licensed in your instrument before they will appear in the Mode menu. Some modes also require the presence of specific hardware.

status bits should be read, and any errors resolved, prior to switching modes. Error conditions that exist prior to switching modes cannot be detected using the condition registers after the mode change. This is true unless they recur after the mode change, although transitions of these conditions can be detected using the event registers.
Changing modes resets all SCPI status registers and mask registers to their power-on defaults. Hence, any event or condition register masks must be re-established after a mode change. Also note that the power up status bit is set by any mode change, since that is the default state after power up.
If Preset Type Mode is selected, then the analyzer settings are preset but it stays in that selected mode.
:INSTrument:NSELect
:INSTrument:NSELect?
Remote Command Notes: Enter one of the following integers in the command to set the analyzer mode.
EDGE with GSM 13 EDGEGSM
89600 VSA Link Software 231 LINK
Noise Figure 219 NFIGURE
Phase Noise 14 PNOISE
Spectrum Analysis 1 SA
W-CDMA for 3GPP 9 WCDMA

3.5.3 Application Mode Catalog Query (Remote command only)
Returns a comma separated list of strings that contain the names of all the installed applications/modes. These names can only be used with the INST:SELECT command.
:INSTrument:CATalog?
Example: INST:CAT?

This chapter provides key descriptions and programming information for the front-panel key functions of your analyzer starting with the letters P through Z. The front-panel functions are listed alphabetically and are described with their associated menu keys. The lower-level menu keys are arranged and described as they appear in your analyzer.


Places a marker on the highest peak and displays the search menu. If Peak Search Type (Param) is set to Excursion & Threshold, the peak found must meet the defined peak excursion and threshold values. (See “Search Param” on page 215.) Peaks that are less than 1% of the current span away from 0 Hz are ignored. For example, if Span is 1 MHz, peaks will not be found between −10 kHz and +10 kHz. If no valid peak is found, an error (No Peak Found) is displayed. To clear this message, press ESC before attempting another search.
The peak search parameters are Peak Threshold and Peak Excursion. All searches except Peak Search (Next Peak, Next Left, Next Right, Peak Table, SCPI “PEAKS” command) obey the Search Parameters, which means that only peaks which rise above the Peak Threshold by at least the Peak Excursion are found. Peak Search obeys then when Peak Search is in PARAM mode. When the Peak Search key is in MAX mode, the search parameters are ignored for a Peak Search.
:CALCulate:MARKer[1]|2|3|4:MAXimum
Remote Command Notes: The :CALC:MARK:PEAK:SEARC:MODE MAX|PAR command specifies how a peak is identified for use with the marker commands. See “Peak Search” on page 218.
Example: CALC:MARK2:MAX performs a peak search using marker 2.
CALC:MARK2:Y? queries the marker amplitude (Y-axis) value for marker 2. For more information on this command, see “Marker” on page 177.
CALC:MARK2:X? queries the marker frequency or time (X-axis) value for marker 2. For more information on this command, see “Marker” on page 177.
Places the marker on the next highest peak with an amplitude less than the current peak. The peak must meet the defined peak excursion and threshold values. Peaks that are less than 1% of the current span away from 0 Hz are ignored. If no valid peak is found, an error (No Peak Found) is displayed. Press ESC to clear this message before attempting another search. (Also see the Peak Excursn and Pk Threshold key descriptions.)
State Saved: Not part of saved state.
Peak Search

:CALCulate:MARKer[1]|2|3|4:MAXimum:NEXT
Example: CALC:MARK2:MAX:NEXT selects marker 2 and moves it to the next highest peak.
Moves the marker to the next peak to the right of the current marker. The peak must meet the defined peak excursion and threshold limits. Peaks that are less than 1% of the current span away from 0 Hz are ignored. If no valid peak is found, an error “No Peak Found” is displayed. Press ESC to clear this message before attempting another search. (Also see the Peak Excursn and Pk Threshold key descriptions.)
State Saved: Not part of saved state.
:CALCulate:MARKer[1]|2|3|4:MAXimum:RIGHt
Example: CALC:MARK2:MAX:RIGH selects marker 2 and moves it to the next peak to the right.
Moves the marker to the next peak to the left of the current marker. The peak must meet the defined peak excursion and threshold limits. Peaks that are less than 1% of the current span away from 0 Hz are ignored. If no valid peak is found, an error “No Peak Found” is displayed. Press ESC to clear this message before attempting another search. (Also see the Peak Excursn and Pk Threshold key descriptions.)
State Saved: Not part of saved state.
:CALCulate:MARKer[1]|2|3|4:MAXimum:LEFT
Example: CALC:MARK2:MAX:LEFT selects marker 2 and moves it to the next peak to the left.
Moves the active marker to the minimum detected amplitude value on the current trace.
Peak Search

State Saved: Not part of saved state.
:CALCulate:MARKer[1]|2|3|4:MINimum
Example: CALC:MARK:MIN selects marker 1 and moves it to the minimum amplitude value.
Peak Search

Finds and displays the amplitude and frequency (or time, if in zero span) differences between the highest and lowest trace points by setting a reference marker on the peak signal and placing a ∆marker on the minimum signal.
State Saved: Not part of saved state.
:CALCulate:MARKer[1]|2|3|4:PTPeak
CALC:MARK:Y? queries the delta amplitude value for marker 1. For more information on this command, see “Marker” on page 177.
See “Mkr->CF” on page 197 for the command to select this function.
When a marker is placed on a signal and Continuous Pk is pressed, the marker will remain on the signal even if the signal frequency changes, as long as the amplitude of the signal does not change by more than 3 dB from one sweep to another.
If the signal is lost, an attempt will be made to find it again and maintain the marker on the signal peak. If there are other signals on screen near the same amplitude, one of them may be found instead. Signals near 0 Hz cannot be maintained effectively, because they cannot be distinguished from the LO feedthrough, which is excluded by intent from the search algorithm.
State Saved: Saved in instrument state.
:CALCulate:MARKer[1]|2|3|4:CPEak[:STATe] OFF|ON|0|1
Peak Search

:CALCulate:MARKer[1]|2|3|4:CPEak[:STATe]?
Remote Command Notes: This command may not be used to activate a given marker.
Activates the N dB function. Pressing N dB Points (On) turns on the N dB feature and activates two arrows that are N dB down from the marker. The frequency difference between the two arrows will be displayed in the upper right-hand corner of the display. If the feature is unable to find data N dB below the marker, the value of –100 Hz will be displayed in the upper right-hand corner of the display.
For example, N dB Points can be used to measure the 3 dB bandwidth of a filter in a transmission test with the tracking generator.
The measured signal cannot have more than one peak that is greater than or equal to N dB. A signal must be greater than the peak excursion above the peak threshold to be identified. The setting for peak excursion may be increased from the –6 dB default value so that noise will not be identified as signals. Increasing the value too much may cause a smaller signal to be missed or misinterpreted as part of a larger signal. The amplitude scale may be either linear or logarithmic.
The N dB function follows the active marker. If you turn on a marker after N dB Points has been activated, the arrows will follow that marker. If the marker associated with N dB Points moves, the arrows will move with the marker unless there is no data N dB below the marker.
State Saved: If On, the fact is saved in instrument state.
Peak Search

:CALCulate:BANDwidth:NDB
:CALCulate:BANDwidth:NDB?
:CALCulate:BANDwidth[:STATe] ON|OFF|1|0
:CALCulate:BANDwidth[:STATe]?
Peak Search

Displays the search parameter criteria menu that enables you to adjust the parameters for the peak search functions.
There is no remote command for this key.
4.1.9.1 Peak Excursn
Sets the minimum amplitude variation of signals that the marker can identify as a separate peak. For example, if a value of 10 dB is selected, the marker Next Peak function moves only to peaks that rise more than 10 dB above the Peak Threshold and then fall back to the Peak Threshold. This criteria applies to all traces. This function applies to Next Peak, Next Peak Left, and Next Peak Right. If Peak Search is set to Param, it also applies to Peak Search.
Dependencies/ Couplings: This function is not available when Y-Axis is set to Frequency instead of Amplitude.
State Saved: Saved in instrument state.
:CALCulate:MARKer[1]|2|3|4:PEAK:EXCursion UP|DOWN|MIN|MAX|DEFault
:CALCulate:MARKer[1]|2|3|4:PEAK:EXCursion?
Peak Search

you want Peak Search to take advantage of the threshold excursion.
Example: :CALC:MARK:PEAK:EXC 30 DB sets the minimum peak excursion requirement to 30 dB.
See the full example for the CALC:MARK:PEAK:SEAR:MODE command below.
4.1.9.2 Pk Threshold
Specifies the minimum signal level for the analyzer internal peak identification routine to recognize as a peak. A signal must rise above the Peak Threshold by the value specified in Peak Excursn, then fall back to the Peak Threshold, to be considered a peak. This applies to all traces and all windows. Press ESC or select another active function to hide the threshold line. Applies to Next Peak, Next Peak Left, and Next Peak Right. If Peak Search is set to Param, it also applies to Peak Search.
Dependencies/ Couplings: This function is not available when Y-Axis is set to Frequency instead of Amplitude.
State Saved: Saved in instrument state.
Knob Increment: amp scale = LOG: 1 % of dB/div or 0.01 (the larger of the two)
amp scale = LIN:
:CALCulate:MARKer[1]|2|3|4:PEAK:THReshold UP|DOWN|MIN|MAX|DEFault
:CALCulate:MARKer[1]|2|3|4:PEAK:THReshold?
Remote Command Notes: CALC:MARK:PEAK:SEAR:MODE must be set to PARameter if you want peak search to take advantage of the threshold excursion.
Peak Search

Example: :CALC:MARK:PEAK:THR -60 dBm sets the threshold to -60 dBm.
See the full example for the CALC:MARK:PEAK:SEAR:MODE command below.
Peak Search

4.1.9.3 Peak Search
Sets the mode for Peak Search to either Max or Param and applies to Peak Search only.
State Saved: Saved in instrument state.
Factory Default: Maximum.
:CALCulate:MARKer:PEAK[1]|2|3|4:SEARch:MODE PARameter|MAXimum
:CALCulate:MARKer:PEAK[1]|2|3|4:SEARch:MODE?
Remote Command Notes: If mode is set to MAXimum (Max), peak search will place the marker at the maximum amplitude in the trace. If mode is set to PARameter (Param), peak search will place the marker at the highest peak that rises and falls by at least the peak excursion above the peak threshold. If no peak meets the excursion and threshold criteria, a No Peak Found error (error 202) is issued.
The following commands are not affected by the setting of CALC:MARK:PEAK SEAR:MODE. They will always use the parameter search mode that defines peaks based on peak excursion and peak threshold.
:CALCulate:MARKer[1]|2|3|4:MAXimum
:CALCulate:MARKer[1]|2|3|4:MAXimum:LEFT
:CALCulate:MARKer[1]|2|3|4:MAXimum:NEXT
:CALCulate:MARKer[1]|2|3|4:MAXimum:RIGHt
:CALCulate:MARKer:PEAK:TABLe:STATe OFF|ON|0|1
Example: CALC:MARK:PEAK:SEARC:MODE PAR sets the parameter search mode.
CALC:MARK:PEAK:THR –60 dBm sets the threshold to –60 dBm.
CALC:MARK:PEAK:EXC 30 dB sets the minimum peak excursion requirement to 30 dB.
CALC:MARK:STAT ON turns on marker number 1 and puts it on the active trace at mid screen.
Peak Search

CALC:MARK:MAX puts marker 1 on the highest peak that is at least 30 dB above the –60 dBm threshold.
:CALC:MARK:Y? returns the y-axis (amplitude) value of the marker in current y-axis units.
Peak Search

Peak Search

Presetting the instrument provides a known convenient starting point of the instrument state for making measurements. There are three possible actions when you press the Preset key:
— Pressing User Preset resets to the settings/values that you have previously defined as
the User preset state using the Save User Preset key.
— Pressing Mode Preset does not change the mode; it only resets the current mode
settings to the factory defaults.
— Pressing Factory Preset resets the settings for all the modes to the factory defaults.
The factory preset mode is Spectrum Analysis with continuous sweep. If you are not already in the Spectrum Analysis mode, it switches to that mode.
None of these instrument presets resets “persistent” functions such as GPIB address, time/date display style, or auto-alignment state to the factory defaults. See “Restore Sys Defaults” on page 231.
If Preset Type is set to Factory in the System menu, pressing the Preset front-panel key performs a factory preset, which implements the following:
Preset

Dependencies/ Couplings: Depends on the preset type (user, mode or factory) setting in the System, Power On/Preset keys.
SCPI Status Bits/ OPC Dependencies: Clears all pending OPC bits. The status byte is set to 0.
:SYSTem:PRESet
Remote Command Notes: The SYSTem:PRESet command immediately presets the instrument state to values dependent on the preset type that is currently selected (FACTory, USER, MODE).
SYSTem:PRESet will not reset “persistent” functions such as IP address, time/date display style, or auto-alignment state to their factory defaults. Use SYSTem:PRESet:PERSistent. See “Restore Sys Defaults” on page 231.
SYSTem:PRESet:TYPE sets the type of preset. See “Preset Type” on page 211.
Example: :SYST:PRES:TYPE MODE sets the preset mode type to mode. See “Preset Type” on page 211.
:SYST:PRES presets the instrument to the currently selected preset type.
This key is only available when the preset type is set to User. Press System, Power On/Preset, Preset Type, User.
Restores the analyzer to a user defined state that has been saved for all analyzer modes. You defined this state when Save User Preset was pressed. If the you have never saved a user preset state, then the factory preset state is stored as the user preset state. If the user preset state has been saved but the load fails for any reason, the error message: Unable to load user state is displayed in the status line and the state is reset to whatever it was before the Preset key was pressed. This can sometimes happen if firmware has been upgraded or applications have been (un)installed after the user preset state was saved. Save User Preset can be accessed by pressing System, Power On/Preset.
:SYSTem:PRESet See “:SYSTem:PRESet” on page 186.
SYST:PRES:TYPE USER selects the user type preset
SYST:PRES immediately presets the analyzer to the user preset.
Preset

Preset

This key is only available when the preset type is set to User. Press System, Power On/Preset, Preset Type, User.
A mode preset does not change the mode. It resets only the current mode settings to the factory defaults.
See “:SYSTem:PRESet” on page 186.
The *RST command always does a mode preset, except that it sets the instrument to do a single sweep/measurement instead of continuous.
SYST:PRES immediately presets the current analyzer mode
This key is available only when the preset type is set to User. Press System, Power On/Preset, Preset Type, Factory.
A factory preset returns instrument settings to the factory default values. If you are not already in the spectrum analysis mode, it switches to that mode. A factory preset does not reset “persistent” functions such as GPIB address, time/date display style, or auto-alignment state (see “Restore Sys Defaults” on page 231.)
:SYSTem:PRESet See “:SYSTem:PRESet” on page 186.
Remote Command Notes: This command will not reset “persistent” functions such as GPIB address, time/date display style, or auto-alignment state to their factory defaults. Use SYSTem:PRESet:PERSistent. See “Restore Sys Defaults” on page 231.
SYST:PRES immediately presets the analyzer to its factory defaults
Preset

This key is only available when the Preset Type is set to User. Press System, Power On/Preset, Preset Type, User.
System, Power On/Preset. See “Save User Preset” on page 213
:SYSTem:PRESet[:USER]:SAVE
Preset

Preset

Initiates an output of the display data to the currently defined printer. The screen remains frozen (no further sweeps are taken) until the data transfer to the printer is complete. Refer to the key description for Print Setup on page 229 for more information about the printer functions.
There must be a valid printer set up for the print function to work. The Getting Started Guide includes additional printer installation information.
If you need to abort a print in progress, use the Esc (escape) key.
:HCOPy[:IMMediate]
This command aborts the print that is currently in process.
:HCOPy:ABORt


Displays the functions that specify a particular printer and control its output.
There is no remote command for this key.
Enables you to define a printer by selecting its printer language and color capability.
Supported printers are equipped with a parallel interface. (A supported printer is one that accepts Printer Control Language Level 3 or 5). Your printer language can be found in its documentation or in the specifications found on the manufacturer’s web page.
The table below lists some current Hewlett-Packard printers and their settings.
Printer Models Language Type Color Capable
HP DeskJet 310 PCL3 yes
HP DeskJet 320 PCL3 yes
HP DeskJet 400 PCL3 yes
HP DeskJet 670C, 672C, 680C, 682C PCL3 yes
HP DeskJet 720C, 722C Windows only (not compatible)
HP DeskJet 600C, 660C, 670C, 680C, 690C PCL3 yes
HP DeskJet 820C Windows only (not compatible)
HP DeskJet 840C, 850C, 870C, 890C, 895C PCL3 yes
HP DeskJet 935C, 990C PCL3 yes
HP DeskJet 1120C PCL3 yes
HP LaserJet 4L, 4P PCL5 no
HP LaserJet 5, 5L, 5M, 5P, 5MP, 5N PCL5 no
HP LaserJet 6, 6L, 6M, 6P, 6MP PCL5 no
HP Professional Series 2500CM PCL3 yes
HP DesignJet 755CM PCL5 yes
Print Setup

There is no remote command for this key.
4.4.1.1 Language
Lets you define your printer language as a PCL3 (Deskjet) or PCL5 (Laserjet) printer.
State Saved: Persistent, survives Preset and power cycle, but not saved in Instrument State.
:HCOPy:DEVice:LANGuage PCL3|PCL5
:HCOPy:DEVice:LANGuage?
4.4.1.2 Color Capable
Allows you to define whether you printer is color capable (Yes) or not (No).
State Saved: Persistent, survives Preset and power cycle, but not saved in Instrument State.
:HCOPy:DEVice:COLor NO|YES
:HCOPy:DEVice:COLor?
Print Setup

Allows you to select either Portrait or Landscape printing. Landscape is not available with a PCL3 (Deskjet) printer.
State Saved: Persistent, survives Preset and power cycle, but not saved in Instrument State.
:HCOPy:PAGE:ORIentation LANDscape|PORTrait
:HCOPy:PAGE:ORIentation?
4.4.2.1 Portrait
Selects Portrait orientation for the printouts from the analyzer.
See “Orientation” on page 231.
4.4.2.2 Landscape
Selects Landscape orientation for the printouts from the analyzer.
See “Orientation” on page 231.
Print Setup

Selects the number of display prints per page when orientation is set to Portrait. The page will be ejected after the selected number of prints has been printed.
State Saved: Persistent, survives Preset and power cycle, but not saved in Instrument State.
:HCOPy:PAGE:PRINts
:HCOPy:PAGE:PRINts?
Ejects your printed page.
:HCOPy:ITEM:FFEed[:IMMediate]
Ejects the page if prints per page is set to 2 and only 1 print has completed. Otherwise the page automatically ejects after the print is complete.
Allows you to select from the following page sizes: Executive, Letter, Legal, Ledger, A4, and A3.
State Saved: Persistent, survives Preset and power cycle, but not saved in Instrument State.
:HCOPy:PAGE:SIZE A|B|A3|A4|LETTer|LEGal|EXECutive|LEDGer
Print Setup

:HCOPy:PAGE:SIZE?
Remote Command Notes: Page size “A” is letter, and page size “B” is ledger. There is no size standardization for “legal” or “executive.”
Print Setup

4.4.5.1 Executive, Letter, Legal, Ledger, A4, or A3
Selectable page sizes available are as follows: Executive, Letter, Legal, Ledger, A4, and A3.
State Saved: Persistent, survives Preset and power cycle, but not saved in Instrument State.
:HCOPy:IMAGe:COLor[:STATe] OFF|ON|0|1
:HCOPy:IMAGe:COLor[:STATe]?
Allows you to select between color or black and white printing on color-capable printers. This key is inactive (greyed out) if Color Capable is set to No, see page 230.
State Saved: Persistent, survives Preset and power cycle, but not saved in Instrument State.
:HCOPy:IMAGe:COLor[:STATe] OFF|ON|0|1
:HCOPy:IMAGe:COLor[:STATe]?
Print Setup

Saves analyzer states, traces, and screen data to a floppy (A:) drive or internal flash memory (C:) drive, as configured by the File menu. For example, if you have configured the instrument to save a trace to the C: drive, every time you press Save, it will save the current trace to a file with a new default trace file name.
You must first configure the save file Type, Format, Source, and Destination by using File, Save before pressing the front-panel Save key. Pressing the front-panel Save key will then be the same as pressing File, Save, Save Now.
See “File” on page 127.
Use :MMEMory:STORe:SCReen
Use :MMEMory:STORe:STATe 1,
Use :MMEMory:STORe:TRACe ,
Use :MMEMory:STORe:LIMit LLINE1|LLINE2,
Use :MMEMory:STORe:CORRection ANTenna|CABLe|OTHer|USER,
Save

Save

SPAN / X Scale
Activates the Span function and displays the menu of span functions.
Changes the displayed frequency range symmetrically about the center frequency. Setting the span to 0 Hz puts the analyzer into zero span.
Dependencies/ Couplings: Span is coupled to center frequency. The maximum span may be limited by the center frequency setting.
Factory Preset:
E4440A 26.49 GHz
E4443A 6.69 GHz
E4445A 13.19 GHz
E4446A 43.99 GHz
E4448A 49.99 GHz
Range:
Model 0 Hz, Span Range
E4440A 10 Hz to 26.5 GHz
E4443A 10 Hz to 6.7 GHz
E4445A 10 Hz to 13.2 GHz
E4446A 10 Hz to 44.0 GHz
E4448A 10 Hz to 50.0 GHz
Model Span

SPAN / X Scale
[:SENSe]:FREQuency:SPAN
[:SENSe]:FREQuency:SPAN?
Turns on signal tracking and activates the span function. Entering a new span value will then change the span while keeping the marker used for signal tracking centered on the screen. Pressing Span Zoom is the same as pressing: Frequency, Signal Track (On), and Span. See “Signal Track” on page 168.
No equivalent SCPI command.
Changes the displayed frequency span to show the full frequency range of the analyzer. When using external mixing, it changes the displayed frequency span to the frequency range specified for the selected external mixing band.
[:SENSe]:FREQuency:SPAN:FULL
Changes the displayed frequency span to zero Hertz. The horizontal axis changes to time rather than frequency. The input signal that is at the current center frequency is the displayed amplitude. This is a special operation mode that changes several measurement functions/couplings. The instrument behavior is similar to an oscilloscope with a frequency selective detector installed in front of the oscilloscope. See Application Note 150 for more information on how to use this mode.

SPAN / X Scale
See [:SENSe]:FREQuency:SPAN 0 Hz
Changes the displayed frequency span to the previous span setting. If it is pressed immediately after Signal Track or Span Zoom is turned off, then span setting returns to the span that was in effect before these function were turned on.
[:SENSe]:FREQuency:SPAN:PREVious

SPAN / X Scale

4.7 SWEEP
Activates the Sweep Time function and displays the sweep function menu keys.
Selects the length of time in which the spectrum analyzer sweeps the displayed frequency span. In swept spans, the sweep time varies from 1 millisecond to 2000 seconds plus time for setup which is not calculated as part of the sweep time. Reducing the sweep time increases the rate of sweeps. In zero span, the sweep time may be set from 1 µs to 6000 s. In FFT spans, the sweep time is not controlled by the user, but is an estimate of the time required to make FFT measurements. Sweep time is coupled to RBW and VBW, so changing those parameters may change the sweep time. When the analyzer has been set to FFT, Sweep Time, and Auto Sweep Time are disabled (greyed out).
You may select a sweep time less than the auto-coupled value; however this may generate measurement errors. If this happens, the error message: Meas Uncal will appear in the upper right corner of the display.
NOTE In zero span or FFT sweeps, the auto/manual function of this key is not applicable. When Sweep Time (Auto) is selected in non-zero span, any changes to Sweep Time while in zero span will revert to the Auto value when you return to non-zero span. When Sweep Time (Man) is selected in non-zero span, any changes to Sweep Time while in zero span will be maintained when you return to non-zero span (within the available sweep time range). When entering zero span from a non-zero span, the sweep time does not change.
Annunciation/ Annotation: The sweep is displayed in the lower-right corner of the screen. Also, the points displayed parenthetically.
E4440A 66.24 ms
E4443A 11.16 ms
E4445A 22.00 ms
E4446A 110.00 ms
E4448A 125.00 ms
Model Sweep Time

[:SENSe]:SWEep:TIME
[:SENSe]:SWEep:TIME?
[:SENSe]:SWEep:TIME:AUTO OFF|ON|0|1
[:SENSe]:SWEep:TIME:AUTO?
4.7.2 Sweep
Switches the analyzer between continuous-sweep and single-sweep mode. Pressing the front-panel Single key will also put the analyzer in single-sweep mode, then take a sweep.
:INITiate:CONTinuous OFF|ON|0|1
:INITiate:CONTinuous?
:INITiate:[IMMediate]
Remote Command Notes: This command affects sweep if not in a measurement, and affects trigger when in a measurement. A “measurement” refers to any of the functions under the MEASURE key. This corresponds to continuous sweep or single sweep operation when not in a measurement, and continuous measurement or single measurement operation when in a measurement.
When NOT in a measurement, this command does the following:

When in a measurement, this command does the following:
Use the :TRIGger[:SEQuence]:SOURce command to select the trigger source.
Switches the analyzer between normal and accuracy sweep states. Provides you the ability to select the rules for the control of sweep time when Sweep Time is set to Auto. Setting Auto Sweep Time to Accy will result in slower sweep times, usually about three times as long, but better amplitude accuracy for CW signals. The instrument specifications only apply when Sweep Time is set to Auto, and Auto Sweep Time is set to Accy.
Additional amplitude errors occur when Auto Sweep Time is set to Norm are usually well under 0.1 dB, though they are not guaranteed. Because of the faster sweep times and still low errors, Norm is the preferred setting of Auto Sweep Time. Also, when the Auto All function is performed, Auto Sweep Time is set to Norm.
Key Notes: Greyed out in Zero Span and FFT, but still shows user selected states (Norm or Accy). Goes to Norm on Auto All.
Dependencies/ Couplings: Key is greyed out in Zero Span and FFT, but still shows user selected states (Norm or Accy). Goes to Norm on Auto All.
[:SENSe]:SWEep:TIME:AUTO:RULes NORMal|ACCuracy
[:SENSe]:SWEep:TIME:AUTO:RULes?

Sets the number of points per sweep, from 101 to 8192 in non-zero span and 2 to 8192 in zero span. Resolution of setting the sweep time will depend on the number of points selected. If Factory Preset is selected, or the analyzer power is cycled, the number of points per sweep will default to 601. The current value of points is displayed parenthetically, next to the sweep time in the lower-right corner of the display (refer to “Display Annotation” in your Getting Started guide).
Changing the number of points has several effects on the analyzer. Since markers are read at the point location, the marker reading may change. All trace data for the active trace is cleared. If sweep is set to Cont (press Sweep, Sweep), a new sweep begins immediately. If average is set to On (press BW/Avg, Average), the averaging starts over with a count of 0. If limit lines are set to On (press Display, Limits, Modify, Limit 1 or 2), the limit lines are updated.
This function is coupled with the span setting. Increasing the span can change the number of sweep points. In order to maintain accurate measurements, the number of sweep points is increased if required to keep the width of each sweep point at most 150 MHz.
When zone span is set to On (press Span, Zone. Zone), each window has its own value for points.
All trace data is erased Any traces in view mode will go to blank mode Sweep time is re-quantized Any limit lines that are on will be updated

[:SENSe]:SWEep:POINts
[:SENSe]:SWEep:POINts?


Displays the System menu keys to control overall System functions. This is also the GPIB “LOCAL” key. Pressing System after the analyzer has been placed in the remote GPIB mode returns it to the local mode and enables front-panel control. During GPIB operation, “R” appears in the upper-right corner of the display indicating the instrument is in Remote mode. A “T”, “L,” or “S” may appear during remote operation, indicating Talk, Listen, or Service request. Pressing the System (Local) key removes the R symbol in the upper-right corner.
There is no remote command for this key.
Accesses a display of the last 30 (?) errors reported. The most recent error will appear at the top of the list. The first error listed will be the first error removed if the error list is longer than 30 entries. If the same error message occurs several times the error message will be incremented rather than added to the list as a new error message. If there is more than one of the same type of error, the date and time identify the first time and the last time an error occurred and the number of identical errors is shown.
Dependencies/ Couplings: Removes any active functions.
Saved State: Not saved in instrument state.

:SYSTem:ERRor[:NEXT]?
Remote Command Notes: The :SYSTem:ERRor[:NEXT]? command queries the earliest entry to the error queue and then deletes that entry.
Example: SYST:ERR? returns ,, for example -113,“Undefined header”.
*CLS clears the entire error queue.
4.8.1.1 Previous Page
Displays the previous page of the Show Errors screen. This key is inactive (greyed out) if there is no previous page.
4.8.1.2 Next Page
Displays the next page of the Show Errors screen. This key is inactive (greyed out) if there is no next page.
4.8.1.3 Clear Error Queue
Clears the front-panel error queue from the Show Errors display.
4.8.2 Power On/Preset
Displays keys that enable you to define the instrument power-on state and user preset state.
NOTE If Power On is set to Preset, and Preset Type is set to Factory, turning on the analyzer performs a factory preset. The last state of the analyzer (before it was turned off) is recalled if Power On is set to Last. The user preset state is recalled if Power On is set to Preset and Preset Type is set to User. (However, if there is no user preset state saved, the analyzer performs a factory preset.)
System

4.8.2.1 Power On
Enables you to set the state the analyzer will be put in when it is powered on. You can set the state to Last or Preset.
If Power On is set to Last, then the instrument returns to the last saved instrument state and all the modes are restored to that last state. The instrument saves its current state internally every 2 minutes. So the last saved state, that is restored at power-on, is the state at power-off or within a maximum of two minutes before power-off.
If Power On is set to Preset, the instrument state is determined by the preset type setting prior to turning the power off.
The setting (Last or Preset) of the Power On function is not changed by pressing Preset. Use the Power On/Preset menu key function to change the setting of the analyzer state that is recalled at power on. Limit lines are not recalled when the analyzer is powered on. Refer to “Preset” on page 221 for more information.
State Saved: Survives preset and power cycle, but not saved in Instrument State.
:SYSTem:PON:TYPE PRESet|LAST
:SYSTem:PON:TYPE?
Example: SYST:PON:TYPE LAST defines the power on type as the last state the analyzer was in before power was turned off.
System

4.8.2.2 Preset Type
Enables you to select what type of preset will be initiated when you press the green Preset key or send the remote command, using SYST:PRES. Choose between Factory, Mode, or User defined presets.
State Saved: Survives Preset and power cycle, but is not saved in Instrument State.
Preset: Does not change the Preset Type.
:SYSTem:PRESet:TYPE FACTory|USER|MODE
:SYSTem:PRESet:TYPE?
Remote Command Notes: SYSTem:PRESet:USER:SAVE defines the user preset.
Example: SYST:PRES:TYPE FACT defines the type of preset as the factory preset.
4.8.2.2.1 User
Sets the preset type to User. When you do a preset, the instrument state that you have defined as the user preset setting will be restored. Use the Save User Preset key to define your user preset settings. Refer to “Preset” on page 221 for more information. If you have not saved a user state, then the instrument will save the power-up state for you to use as a default user preset state.
State Saved: Survives Preset and power cycle, but no saved in Instrument State.
See “Preset Type” on page 250.
System

user state.
SYST:PRES:TYPE USER defines the type of preset as the user preset.
With user preset selected, and a user state saved, use SYST:PRES to do a user preset.
4.8.2.2.2 Mode
Sets the preset type to “Mode.” When you do a preset, the current mode factory default instrument state will be restored. A mode preset does not change the mode. Refer to “Preset” on page 221 for more information.
State Saved: Survives Preset and power cycle, but no saved in Instrument State.
:SYSTem:PRESet:TYPE See “Preset Type” on page 250.
Example: SYST:PRES:TYPE MODE defines the type of preset as the mode preset.
After you have selected mode as the preset type, use SYST:PRES to do a mode preset.
4.8.2.2.3 Factory
Sets the preset type to “Factory.” When you do a preset, all of the factory default instrument state will be restored. A factory preset switches the analyzer to the Spectrum Analysis mode and resets the settings of all the modes to the factory defaults (i.e. Spectrum Analysis Mode with continuous sweep). Refer to “Preset” on page 221 for more information.
State Saved: Survives Preset and power cycle, but not saved in Instrument State.
System · Factory

See “Preset Type” on page 250.
Example: SYST:PRES:TYPE FACT defines the type of preset as the factory preset.
With factory preset selected, use SYST:PRES to do a factory preset.
4.8.2.3 Save User Preset
Saves the current state of the analyzer into the User Preset state registor for recall when the instrument Preset Type is set to User and you perform a preset operation. After you save a state here, you must go to the Preset Type key and select User in order to have this state used as the preset state. Refer to “Preset” on page 221 key description for the default factory-configuration settings.
Preset, if the preset type is set to User. See “Save User Preset” on page 225
:SYSTem:PRESet[:USER]:SAVE
Use SYST:PRES:TYPE USER to set factory preset type to “User.”
Then use SYST:PRES to do the preset.
Displays the Time/Date function menu keys used to set and display the real-time clock.
There is no remote command for this key.
System

4.8.3.1 Time/Date
Turns the display of the real-time clock on or off.
Annunciation/ Annotation: Current time and date are displayed in the upper-left corner of the screen.
State Saved: Survives Preset and power cycle, but not saved in Instrument State.
:DISPlay:ANNotation:CLOCk[:STATe] ON|OFF
:DISPlay:ANNotation:CLOCk[:STATe]?
4.8.3.2 Date Format
Enables you to set the date display to month-day-year or day-month-year. It is set to a month-day-year format when the instrument System Defaults are restored. This key only effects display of date at the top of the screen, not in the file catalog.
State Saved: Survives Preset and power cycle, but not saved in Instrument State.
:DISPlay:ANNotation:CLOCk:DATE:FORMat MDY|DMY
:DISPlay:ANNotation:CLOCk:DATE:FORMat?
4.8.3.3 Set Time
Enables you to set the time of the real-time clock. Enter the time in 24 hour HHMMSS format.
System

State Saved: Survives Preset and power cycle, but not saved in Instrument State.
Range: Hour (HH): 00 to 23.
Minute (MM): 00 to 59.
Second (SS): 00 to 59.
:SYSTem:TIME ,,
:SYSTem:TIME?
Example: SYST:TIME 12,42,00 Sets the clock to 12:42:00 PM.
4.8.3.4 Adjust Time Setting (Remote Command Only)
Adjust the instruments internal time by the value entered.
Terminators: No units are allowed with the command.
:SYSTem:TIME:ADJust
Example: SYST:TIME:ADJ 3600 will advance the time one hour.
SYST:TIME:ADJ -86400 will back the date up one day, without changing the time of day (minutes or seconds).
4.8.3.5 Set Date
Allows you to set the date of the real-time clock. Enter the date in the YYYYMMDD format.
System · DefaultTerminator seconds

restored by System, Restore Sys Defaults.
Month (MM):01 to 12
:SYSTem:DATE ,,
:SYSTem:DATE?
Displays functions that control the automatic alignment of the instrument and load default values for the alignment system.
NOTE Most CALibration commands execute in the background, permitting other SCPI commands to be processed concurrently. If a measurement command is sent right after a CALibration command, there can be interaction between background alignments and the measurement. The *WAI command should be issued after any CALibration command and before the measurement command. Note that sending the query form of a CAL? command will automatically hold off any following commands until the query value is returned.
Bit 0 (CALibration) must be set to 1 in the STATus:OPERation:ENABle register to ensure that the *WAI command waits for calibration to complete. This is the factory preset and *RST setting.
There is no remote command for this key.
4.8.4.1 Auto Align
Allows you to turn the instrument automatic alignment On or Off. or select Alert to be
System

alerted that alignments are needed.
System

* There are 2 very quick alignments, invisible to the user, that are done every few minutes or when certain settings are changed. These still occur, even if AutoAlign is set to Off. These alignments are the Current SysGain and Current IF Flatness methods which can also be forced to occur by user under the Align Subsys menu.
State Saved: Survives Preset and power cycle, but not saved in Instrument State.
Factory Preset: Not affected by Preset.
:CALibration:AUTO OFF|ON|ALERt
:CALibration:AUTO?
4.8.4.2 Align All Now
Accesses the Align Now menu keys and immediately executes an alignment cycle of all the subsystems (Align RF, Align IF, Align ADC, and Align Current Sys Gain). The instrument will stop any measurement currently underway, perform the full alignment, then restart the measurement from the beginning (similar to pressing the Restart key). All other operations are stopped and the alignments will be visible on the display.
:CALibration[:ALL] Performs a full alignment.
The following three commands perform a full alignment and return a number indicating the success of the alignment. A zero is returned if the alignment is successful. A one is returned if any part of the alignment fails.
:CALibration[:ALL]?
*TST?
Example: CAL? The query performs a full alignment and returns a number indicating the success of the alignment. A zero is returned if the alignment is successful.
4.8.4.3 Frequency Corrections (Remote Command Only)]
Turns the internal frequency corrections on/off.
System

:CALibration:FREQuency[:STATe] OFF|ON|0|1
:CALibration:FREQuency[:STATe]?
4.8.4.4 Align Subsys
Displays keys that enable you to activate a partial alignment.
There is no remote command for this key.
4.8.4.4.1 Align RF
Initiates an alignment on the RF assembly.
Annunciation/ Annotation: Will display a series of pop-up message boxes, indicating an alignment is in progress.
If a signal is present which interfere with the alignment a message Align RF 50 MHz signal detected. will be displayed.
:CALibration:RF
:CALibration:RF?
Remote Command Notes: The query performs the alignment and returns a zero if the alignment is successful.
Example: CAL:RF?
4.8.4.4.2 Align IF
System · fails

Initiates an alignment on the IF assembly.
Annunciation/ Annotation: Will display a series of pop-up message boxes indicating alignment progress.
SCPI Status Bits/ OPC Dependencies: Status questionable calibration Bit 4 will be set if Align RF fails.
:CALibration:IF
:CALibration:IF?
Remote Command Notes: The query performs the alignment and returns a zero if the alignment is successful.
Example: CAL:IF?
4.8.4.4.3 Align ADC
Initiates an alignment on the ADC circuitry.
Annunciation/ Annotation: Message boxes indicating alignment of ADC progress.
SCPI Status Bits/ OPC Dependencies: Status questionable calibration Bit 6 will be set if Align ADC fails.
:CALibration:ADC
:CALibration:ADC?
Remote Command Notes: The query performs the alignment and returns a zero if the alignment is successful.
Example: CAL:ADC?
4.8.4.4.4 Align Current IF Flatness
Initiates an alignment of the current IF flatness, for the purpose of improving absolute amplitude within FFT Sweeps and improving group delay in some digital demodulation measurements.
System

aligned (a subset of the Align All Now message boxes).
:CALibration:FLATness:IF
:CALibration:FLATness:IF?
Remote Command Notes: The query performs the alignment and returns a zero if the alignment is successful.
Example: CAL:FLAT:IF?
System

4.8.4.4.5 Align Current SysGain
Initiates a fine-tuning adjustment of the system gain, primarily to correct for small amplitude variations that occur as resolution BW is switched.
Annunciation/ Annotation: Will display a series of pop-up message boxes explaining what is being aligned (a subset of the Align All Now message boxes).
:CALibration:GAIN:CSYStem
:CALibration:GAIN:CSYStem?
Remote Command Notes: The query performs the alignment and returns a zero if the alignment is successful.
Example: CAL:GAIN:CSYS?
4.8.4.5 Restore Align Defaults
Loads the default values for the alignment system, turns on the frequency corrections, and resets the timebase to the factory values. Align All Now must be executed 3 times after pressing Restore Align Defaults to meet specifications.
Annunciation/ Annotation: System Alignments, Align Now All required.
SCPI Status Bits/ OPC Dependencies: Status questionable calibration Bit 14 will be set.
:CALibration:DATA:DEFault
4.8.4.6 Set Calibration Display Detail (Remote Command Only)
Controls the amount of detail shown on the display while the alignment routines are running. The routines run faster if the display level is off, so they do not have to update the display.
Off - displays no trace points
Low - displays every 10th trace
System

High - displays every trace
:CALibration:DISPlay:LEVel OFF|LOW|HIGH
:CALibration:DISPlay:LEVel?
4.8.4.7 Select Time Corrections (Remote Command Only)
Controls time corrections used to compensate for the complex (magnitude and phase) response of the analog and digital IF hardware. When only scalar (magnitude) FFT flatness is required, time corrections take more CPU cycles and so are less efficient than frequency corrections. For demod or other time-based (not FFT) measurements, only time corrections can improve the flatness that results from imperfect IF hardware. When the time correction functionality is set to Auto (the default), the individual measurements activate the corrections when they are needed.
Always return time corrections to Auto.
:CALibration:TCORrections AUTO|ON|OFF
Displays the keys and menus that enable you to identify and change the current GPIB address and LAN settings.
There is no remote command for this key.
System

4.8.5.1 GPIB Address
Shows the current GPIB address and allows you to change this value using the numeric keyboard. The new value is displayed in the active function area. The GPIB port is always active. The knob and step keys are not active for this function.
State Saved: Survives Preset and power cycle, but not saved in Instrument State.
:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess
:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess?
4.8.5.2 IP Address
Allows you to set the IP (internet protocol) address, domain name and node (host) name for the instrument. The IP address of the instrument can be changed by entering a numeric address composed of numbers and decimal points. Press ENTER to complete the entry.
State Saved: Survives Preset and power cycle, but not saved in Instrument State.
:SYSTem:COMMunicate:LAN[:SELF]:IP
:SYSTem:COMMunicate:LAN[:SELF]:IP?
System

Sets the IP address to 150.222.50.52 and sets the host name to mypsa.
4.8.5.3 Host Name
Displays the host name of the instrument. Pressing the key activates the alpha editor, which enables you to change the host name. (Press ENTER to complete the entry.)
State Saved: Survives Preset and power cycle, but not saved in Instrument State.
See above command :SYSTem:COMMunicate:LAN[:SELF]:IP
Sets the IP address to 150.222.50.52 and sets the host name to mypsa.
4.8.5.4 Host ID (Remote Command Only)
Enables you to query the host ID remotely. The current value of the host ID can be viewed on the display by pressing System, Show System.
Factory Default: Value is unique to your instrument (Not reset by System, Restore Sys Defaults.)
:SYSTem:HID?
Remote Command Notes: The host ID cannot be set remotely, it can only be queried.
Example: SYST:HID?
System

4.8.5.5 Subnet Mask
Changes the subnet mask of the instrument. The subnet mask is a 32-bit address mask used in IP networks to indicate the bits of an IP address that are used for the subnet address. The default address is 255.255.0.0 for a class B network
State Saved: Survives Preset and power cycle, but not saved in Instrument State.
History: Added with firmware revision A.03.00.
There is no equivalent remote command.
4.8.5.6 Gateway Address
Allows you to set the gateway address. The gateway address feature is used to manipulate the gateway used to reach the destination. The gateway address can be changed by entering a numeric address composed of numbers and decimal points. Press ENTER to complete the entry.
State Saved: Survives Preset and power cycle, but not saved in Instrument State.
History: Added with firmware revision A.03.00.
There is no remote command for this key.
Displays keys to enable SCPI functionality over LAN. There are a number of different ways to send SCPI remote commands to the instrument over the LAN. It can be a problem to have multiple users simultaneously accessing the instrument over the LAN. These keys allow you to limit that somewhat by disabling the telnet socket and/or SICL capability.
There is no remote command for this key.
System

4.8.5.7.1 SCPI Telnet
Turns on/off the SCPI LAN telnet capability allowing you to limit SCPI access over LAN via telnet.
State Saved: Survives Preset and power cycle, but not saved in Instrument State.
:SYSTem:COMMunicate:LAN:SCPI:TELNet:ENABle OFF|ON|0|1
:SYSTem:COMMunicate:LAN:SCPI:TELNet:ENABle?
4.8.5.7.2 SCPI Socket
Turns on/off the capability of establishing Socket LAN sessions. This allows you to limit SCPI access over LAN via socket sessions.
State Saved: Survives Preset and power cycle, but not saved in Instrument State.
:SYSTem:COMMunicate:LAN:SCPI:SOCKet:ENABle OFF|ON|0|1
:SYSTem:COMMunicate:LAN:SCPI:SOCKet:ENABle?
System

4.8.5.7.3 SICL Server
Turns on/off the SICL server capability, enabling you to limit SCPI access over LAN via the SICL server. (SICL IEEE 488.2 protocol.)
Table 4-1 SCPI Default Settings
Maximum Connections The maximum number of connections that can be accessed simultaneously
Instrument Name The name (same as the remote SICL address) of your analyzer
Instrument Logical Unit The unique integer assigned to your analyzer when using SICL LAN
Emulated GPIB Name The name (same as the remote SICL address) of the device used when communicating with your analyzer
Emulated GPIB Logical Unit The unique integer assigned to your device when it is being controlled using SICL LAN
Emulated GPIB Address The emulated GPIB address assigned to your transmitter tester when it is a SICL server (the same as your GPIB address)
State Saved: Survives Preset and power cycle, but not saved in Instrument State.
:SYSTem:COMMunicate:LAN:SCPI:SICL:ENABle OFF|ON|0|1
:SYSTem:COMMunicate:LAN:SCPI:SICL:ENABle?

Displays functions that control the external frequency reference.
There is no remote command for this key.
4.8.6.1 Freq Ref
Specifies the frequency reference as being internal or external. If the frequency reference is specified as internal, the frequency of the reference is automatically identified as being 10 MHz. If the frequency reference is specified as external, you must enter the frequency of the external reference being used. If External Reference is selected, Ext Ref will appear on the right side of the display.
The frequency of an external frequency reference is not automatically detected. If an external frequency source is selected, the frequency of the source must be entered.
If Ext is selected, and you press Freq Ref, Ext will remain selected and the Ext reference frequency will become the active function. If Freq Ref is pressed again, Int will become selected (at 10 MHz). The Ext reference frequency is remembered and will be used again if Ext is selected.
If the external reference is missing or out of range, or the frequency reference is unlocked, the message “External reference missing or out of range”, will appear on the display.
State Saved: Not Saved in Instrument State. Neither the external reference frequency nor the state of this function (Int or Ext) are affected by factory preset or power cycle. Reset to the factory default (Int, 10 MHz) by pressing System, Restore Sys Defaults.
[:SENSe]:ROSCillator:SOURce INTernal|EXTernal
[:SENSe]:ROSCillator:SOURce?
[:SENSe]:ROSCillator:EXTernal:FREQuency
[:SENSe]:ROSCillator:EXTernal:FREQuency?
Example: Before switching to the external reference source, specify the frequency of the external reference that you plan to use.
ROSC:EXT:FREQ 20 MHz sets the external reference frequency to 20 MHz,
System

but does not select the external reference.
ROSC:SOUR EXT selects the external reference.
System

4.8.6.2 10 MHz Out
Switches the 10 MHz out signal on the rear panel of the analyzer on or off.
State Saved: Not Saved in Instrument State. Not affected by factory preset or power cycle. Reset to the factory default (Off, 10 MHz) by pressing System, Restore Sys Defaults.
[:SENSe]:ROSCillator:OUTPut[:STATe] OFF|ON|0|1
[:SENSe]:ROSCillator:OUTPut[:STATe]?
Displays the number and description of the options installed in your instrument. It also displays the instrument model number, product number, serial number, ethernet address, host ID, firmware revision, revision date, options, and system statistics.
Dependencies/ Couplings: Active function is disabled.
*IDN? Returns four fields separated by commas:
:SYSTem:OPTions?
Returns a string of all the installed instrument options. It is a comma separated list such as: BAC,BAH.
Example: *IDN?
System

4.8.8 Show Hdwr
Gives detailed information about the hardware installed on your instrument.
:SYSTem:CONF:HARD ON|OFF
Displays the Color Palette menu keys that set the display screen attributes.
State Saved: Not saved in Instrument State, survives Preset, and power cycle.
Preset: Not affected by Preset.
There is no remote command for this key.
4.8.9.1 Default
Selects the factory default color palette.
System

4.8.9.2 Vision Impair 1
Selects a special color scheme to accommodate color-deficient vision problems.
There is no remote command for this key.
4.8.9.3 Vision Impair 2
Selects a special color scheme to accommodate color-deficient vision problems.
There is no remote command for this key.
4.8.9.4 Optical Filter
Selects a special color scheme to accommodate protective goggles while viewing lasers.
There is no remote command for this key.
System

System

4.8.9.5 Monochrome
Sets the color palette to single-color mode. The monochrome display uses different shades of green for each green value. This is especially useful for driving external monochrome monitors.
There is no remote command for this key.
Access front-panel diagnostic functions.
There is no remote command for this key.
4.8.10.1 Front Panel Test
Used to test the front-panel keys. It shows a list of all the front panel keys with counter numbers indicating the number times the key is pressed. Press the ESC key to exit the test mode and return to the menu.
Dependencies/ Couplings: The active function is disabled.
Resets the system settings, including most “persistent” functions, to their factory defaults. It also does a Factory Preset that resets the analyzer to the Spectrum Analysis Mode. It does not reset user data such as saved instrument states.
System

Persistent functions are things such as the GPIB address, time/date display style, and auto-alignment state. These are parameters that are unaffected by a power cycle or an instrument preset.
Table 4-2 System Default Settings
Feature Default Setting Restored?
Automatic Alignment On/Alert/Off On yes
Power On Last/Preset Preset yes
Preset Type Factory/User/Mode Mode yes
External Amplifier Gain 0 dBm yes
Input Impedance 50 Ohm yes
GPIB Address 18 yes
IP address 10.10.10.10 no
Host Name mypsa no
Subnet Mask 255.255.0.0 no
SCPI Telnet Port 5023 On yes
SCPI Socket Port 5025 On yes
SICL Server On yes
Instrument Color Palette Default palette yes
Printer Setup, Language PCL3 yes
Printer Setup, Color Capability No yes
Print Orientation Portrait yes
Color Printing Off yes
Prints/ Page 1 yes
Date Format MDY yes
Time Date Display On yes
Verbose (error messages) Off yes
Display Viewing Angle 4 yes
Manual Tracking Adjustment 2048 yes
Page Size Letter yes
Printer Selection Auto yes
System

:SYSTem:PRESet:PERSistent
Accesses the security system to enable licensing for individual options. You can install a measurement mode personality options in an instrument at any time. After you load the personality mode into memory, you must enter a license key to allow access to the option.
For information on how to use this system, see the installation information for individual options in the measurement personalities’ user’s guides.
There is no remote command for this key.
4.8.12.1 Option
Activates the alpha editor enabling you to enter the designation for the option to be installed. An option is a three character string that specifies the option or application that is to be installed, as found in the catalog. To terminate the entry, press Enter or Return. An external keyboard may also be used for this entry. The option number will appear on the second line of the Option key.
State Saved: Not saved in instrument state.
See “License Key” on page 277
System

4.8.12.2 License Key
Activates the alpha editor to allow you to enter the license key number for the option to be installed. The license key number is a hexadecimal number that will require entry of both letters and numbers. Use the front-panel numeric keyboard to enter numerical values. You will see your entry in the active function area. A license key is a 12-character hexadecimal string given with the option. The license key is unique to a specific option installed and instrument host ID. To terminate the entry, press Enter or Return. An external keyboard may also be used for this entry. The license key number will appear on the second line of the License Key menu key.
State Saved: Not saved in instrument state.
Factory Preset: Clears the current value.
:SYSTem:LKEY ,
:SYSTem:LKEY?
Remote Command Notes: The query returns a string that contains the license key for a specified application or option that is already installed in the instrument. The license key will also be returned if the application is not currently in memory, but had been installed at some previous time. The license key is unique to a specific option, host ID and serial number. Host ID can be returned by :SYSTem:HID?.
The query would return “B62A35B37679”
If the instrument does not have a license key for that option, the query would return ““.
4.8.12.3 Activate License
Activates the specified option. When a valid option key number and license key number have been entered, press Activate. If your entry was correct, the option will be enabled and the message: Option activated will appear in the status line of your display. The option number and license key number will also be cleared from the License Key and Option key.
If your entry was incorrect, the error message: License key invalid will appear in the status line of your display. The option number and license key number can be edited until
System

they are correct.
There is no remote command for this key.
4.8.12.4 Delete License
Deletes the license key from memory, however, the option firmware is not deleted.
:SYSTem:LKEY:DELete ,
4.8.12.5 Show License
Displays the number and description of the licenses installed in your instrument.
Dependencies/ Couplings: The active function is disabled.
There is no remote command for this function.
4.8.12.6 Install an Application Mode (Remote command only)
Installs the specified application from an external drive to the instrument. Each application allows you to make a specific set of measurements easily and accurately. Installation requires a 12-character license key that you received with your application. The license key number is unique to the option and instrument serial number. If it cannot be located, contact your local Agilent Technologies and service office to re-obtain the information. (Have the instrument model number, host ID, serial number available.)
System

Example: *OPT? returns a string with all the application options currently installed in the instrument (e.g. “B7J,202,204,BAC”).
Pressing Personality displays information about the personalities installed and their license status, as shown in the following illustration.

Dependencies/ Couplings: The active function is disabled.
These functions are used only for servicing the analyzer. A password is required to access them. Refer to the Service Guide for more information.
System · See also *OPT?

Disables the instrument keyboard to prevent local input when instrument is controlled remotely. An annunciator reading “Klock” alerts the local user that the keyboard is locked. Or you can display a system message using SYSTem:MESSage.
:SYSTem:KLOCK?
Example: SYST:KLOCK?
Enables remote user to send message that will appear in the Status Bar at bottom of the instrument display. New message will overwrite any previous message. Message will remain until removed by use of :SYSTem:MESSage:OFF.
Remarks: Message appears as green text against a black background to differentiate it from internally generated messages which appear as white text against a black background.
The SYSTem:KLOCk command will lock out the front panel keys.
:SYSTem:MESSage
System

Removes any system message from the Status Bar at the bottom of the instrument display. A message can be displayed using the :SYSTem:MESSage command.
:SYSTem:MESSage:OFF
Returns the number of seconds that have elapsed since the instrument was turned on for the very first time.
:SYSTem:PON:ETIMe?
Example: SYST:PON:ETIM?
Returns the SCPI version number with which the instrument complies. The SCPI industry standard changes regularly. This command indicates the version used when creating the instrument SCPI commands.
:SYSTem:VERSion?
Example: SYST:VERS?
System

System

Displays menu keys that enable you to set how trace information is stored and displayed. Each trace is comprised of a series of data points in which x and y axis information is stored. The analyzer updates the information for the active trace with each sweep.
:TRACe[1]|2|3:MODE WRITe|MAXHold|MINHold|VIEW|BLANk
WRITe = Clear Write
MAXHold = Max Hold
MINHold = Min Hold
:TRACe[1]|2|3:MODE?
TRAC:MODE MINH
TRAC:MODE VIEW
TRAC:MODE BLANk
Trace/View · VIEW = View · BLANk = Blank

This query returns the current values of the designated trace amplitude values. The data is terminated with . (For GPIB this is newline, or linefeed, followed by EOI set true. For RS-232 this is newline only.)
The number of trace points returned is set by [:SENSE]:SWE:POIN (from 101 to 8192).
:TRACe[:DATA]?
Remote Command Notes: The FORMat:DATA command describes the different types of data formats that can be used with trace data. See “Format Numeric Data (Remote Command Only)” on page 160.
Use the FORMat:BORDer command to set the byte order. See “Set Data Byte Order (Remote Command Only)” on page 160.
Commands :MMEM:STOR:TRAC and :MMEM:LOAD:TRAC are used to transfer trace data to/from the internal hard drive or floppy drive of the instrument. See “Save Now” on page 137 and “Load Now” on page 144.)
Example: TRAC? TRACE2 queries the analyzer for the contents of trace 2.
Determines which trace the menu keys will affect. Press Trace until the number of the desired trace is underlined.
There is no remote command for this function.
Trace/View · is TRACE1|TRACE2|TRACE3

Erases any data previously stored in the selected trace and continuously displays signals during the sweep of the analyzer.
See “Trace/View” on page 283.
Maintains the maximum level for each trace point of the selected trace (1, 2 or 3), and updates each trace point if a new maximum level is detected in successive sweeps.
See “Trace/View” on page 283.
Maintains the minimum level for each trace point of the selected trace (1, 2 or 3), and updates each trace point if a new minimum level is detected in successive sweeps.
See “Trace/View” on page 283.
Trace/View

Holds and displays the amplitude data of the selected trace. The trace is not updated as the analyzer sweeps.
See “Trace/View” on page 283.
Stores the amplitude data for the selected trace and removes it from the display. The selected trace register will not be updated as the analyzer sweeps.
See “Trace/View” on page 283.
Trace/View

Displays menu keys that enable you to select the trigger mode of a sweep or measurement. When in a trigger mode other than Free Run, the analyzer will begin a sweep only with the proper trigger condition.
In FFT measurements, the trigger controls when the data is acquired for FFT conversion; see “Making Gated FFT Measurements With Your PSA” on page 74.
:TRIGger[:SEQuence]:SOURce IMMediate|VIDeo|LINE|EXTernal[1]|EXTernal2|RFBurst
:TRIGger[:SEQuence]:SOURce?
IMM = Free Run triggering VID = Video–triggers on the video signal level LINE = Line–triggers on the power line signal Ext1 = External Front–enables you to trigger on an externally connected trigger source Ext2 = External Rear–enables you to trigger on an externally connected trigger source
Remote Command Notes: Other trigger-related commands are found in the INITiate and ABORt subsystems.
Sets the trigger to start a new sweep/measurement as soon as the last one has ended (continuous sweep mode) or immediately (single sweep mode).
Dependencies/ Couplings: Trigger Slope and Delay adjustments are not available with Free Run triggering.
See “Trig” on page 287
Trig

Activates the trigger condition that allows the next sweep to start if the detected RF envelope voltage crosses a level set by the video trigger level. When Video is pressed, a line appears on the display. The analyzer triggers when the input signal exceeds the trigger level at the left edge of the display. You can change the trigger level using the step keys, the knob, or the numeric keypad. The line remains as long as video trigger is the trigger type.
Dependencies/ Couplings: Trigger Delay adjustment is not available with Video triggering.
Video triggering is not available when the detector is Average. Marker Functions that set the detector to average (such as Marker Noise or Band/Intvl Power) are not available when the video trigger is on.
This function is not available when the Resolution Bandwidth is less than 1 kHz. If a Resolution Bandwidth less than 1 kHz is selected while in Video Trigger mode, the Trigger mode changes to Free Run.
Range: Using logarithmic scale: from 10 display divisions below the reference level, up to the reference level
For more information, see “Scale Type” on page 52.
See “Trig” on page 287 for the command that sets trigger mode. The following commands set/read the trigger level.
:TRIGger[:SEQuence]:VIDeo:LEVel
:TRIGger[:SEQuence]:VIDeo:LEVel?
:TRIGger[:SEQuence]:VIDeo:LEVel:FREQuency sets the Video Trigger level when in FM Demod, and Demod View is on
:TRIGger[:SEQuence]:VIDeo:LEVel:FREQuency?
Example: TRIG:SOUR VID selects video triggering.
Sets the trigger to start a new sweep/measurement to be synchronized with the next cycle of the line voltage.
Trig

See “Trig” on page 287
Example: TRIG:SOUR LINE selects line triggering.
Sets the trigger to start a new sweep/measurement whenever the external voltage (connected to EXT TRIGGER INPUT on the front panel) passes through approximately 1.5 volts. The external trigger signal must be a 0 V to +5 V TTL-type signal.
SCPI Status Bits/ OPC Dependencies: Line trigger is not available when operating from a dc power source.
See “Trig” on page 287
Example: TRIG:SOUR EXT to select front panel external triggering.
Sets the trigger to start a new sweep/measurement whenever the external voltage (connected to TRIGGER IN on the rear panel) passes through approximately 1.5 volts. The external trigger signal must be a 0 V to +5 V TTL-type signal.
See “Trig” on page 287
Example: TRIG:SOUR EXT2 selects rear panel external triggering.
Allows the analyzer to be triggered by an RF burst envelope signal.
See “Trig” on page 287
Trig

Controls the trigger polarity. It is positive to trigger on a rising edge and negative to trigger on a falling edge.
Dependencies/ Couplings: Not available for Free Run.
State Saved: Saved in instrument state.
:TRIGger[:SEQuence]:SLOPe POSitive|NEGative
:TRIGger[:SEQuence]:SLOPe?
Allows you to control a time delay during which the analyzer will wait to begin a sweep after receiving an external or line trigger signal. You can use negative delay to pre-trigger the instrument.
Dependencies/ Couplings: This function is not available when Trigger is Free Run or Video.
This function is not available when Gate is on.
State Saved: Saved in instrument state.
:TRIGger[:SEQuence]:DELay
:TRIGger[:SEQuence]:DELay?
:TRIGger[:SEQuence]:DELay:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:DELay:STATe?
Trig

TRIG:DEL 100 ms
This command sets the trigger offset. Trigger offset refers to the specified time interval before or after the trigger event from which data is to be written to the trace, and then displayed. Ordinarily, the trigger offset value is zero, and trace data is displayed beginning at the trigger event. A negative trigger offset value results in the display of trace data prior to the trigger event. A positive trigger offset value results in an effective delay in the display of trace data after the trigger event.
The trigger offset value used when the feature is enabled will depend on the following parameters:
The effective trigger offset value will be re-calculated whenever any of these parameters change.
State Saved: Saved in instrument state.
Range: Hardware specific; dependent upon the ADC being used, current state and the number of sweep points.
:TRIGger[:SEQuence]:OFFSet
:TRIGger[:SEQuence]:OFFSet?
:TRIGger[:SEQuence]:OFFSet:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:OFFSet:STATe?
Remote Command Notes: Trigger offset can only be turned on when in zero span and the resolution bandwidth is 1 kHz or greater (non-digital bandwidths). Trigger offset is available for all trigger modes.
TRIG:OFFS:STAT ON turns on the trigger offset.
Trig

Trig

This chapter provides you with functional descriptions, and remote programming commands for all measurements available to you when you press the front-panel Measure key (Spectrum Analysis mode is selected).

This section provides key descriptions and programming information for each key associated with the measurements available when you press the front-panel Measure key. Note that the information for keys common to all measurements (Measure, Meas Control, Mode Setup, Restart, and Single) is presented first and the information unique to each measurement is grouped by measurement type. Use the table below to locate the key descriptions and remote commands for the measurement of interest.
Key Function Page Location
CCDF Measurement (Power Stat)
Span X Scale 389

Key Function Page Location
Spectrum Emission Mask Measurement
Span X Scale 481



MEASURE (Spectrum Analysis Mode)
In the Spectrum Analysis mode (see the Mode key), this key displays a menu that lets you to make transmitter power measurements such as adjacent channel power, occupied bandwidth, and harmonic distortion measurements. If other modes are available and have been selected, the measurements for that particular mode will be displayed. Some common settings can be made for these measurements using the function under the Mode Setup key. For example, you may select one of several radio standards available by pressing Mode Setup, Radio Std.
Dependencies/ Couplings: When a measurement is activated, Signal Track, Video Averaging, Segmented Sweep, Band Power and Marker Noise will be turned off.
Remote Command Info:
Figure 1 Measurement Group of Commands


MEASURE (Spectrum Analysis Mode)
Measure Commands:
:MEASure:[n]?
This is a fast single-command way to make a measurement using the factory default instrument settings. These are the settings and units that conform to the Mode Setup settings (e.g. radio standard) that you have currently selected.
The scalar measurement results will be returned if the optional [n] value is not included, or is set to 1. If the [n] value is set to a value other than 1, the selected trace data results will be returned. See each command for details of what types of scalar results or trace data results are available.
ASCII is the default format for the data output. (Older versions of Spectrum Analysis and Phase Noise mode measurements only use ASCII.) The binary data formats should be used for handling large blocks of data since they are smaller and faster than the ASCII format. Refer to the FORMat:DATA command for more information.
If you need to change some of the measurement parameters from the factory default settings you can set up the measurement with the CONFigure command. Use the commands in the SENSe: and CALCulate: subsystems to change the settings. Then you can use the READ? command to initiate the measurement and query the results. See Figure 1.
If you need to repeatedly make a given measurement with settings other than the factory defaults, you can use the commands in the SENSe: and CALCulate: subsystems to set up the measurement. Then use the READ? command to initiate the measurement and query results.
Measurement settings persist if you initiate a different measurement and then return to a previous one. Use READ:? if you want to use those persistent settings. If you want to go back to the default settings, use MEASure:?.
Configure Commands:
:CONFigure:<measurement>
This command stops the current measurement (if any) and sets up the instrument for the specified measurement using the factory default instrument settings. It sets the instrument to single measurement mode but should not initiate the taking of measurement data unless INIT:CONTinuous is ON. After you change any measurement settings, the READ command can be used to initiate a measurement without changing the settings back to their defaults.
The CONFigure? query returns the current measurement name.

MEASURE (Spectrum Analysis Mode)
Fetch Commands:
:FETCh:<measurement>[n]?
This command puts selected data from the most recent measurement into the output buffer. Use FETCh if you have already made a good measurement and you want to return several types of data (different [n] values, e.g. both scalars and trace data) from a single measurement. FETCh saves you the time of re-making the measurement. You can only FETCh results from the measurement that is currently active, it will not change to a different measurement.
If you need to get new measurement data, use the READ command, which is equivalent to an INITiate followed by a FETCh.
The scalar measurement results will be returned if the optional [n] value is not included, or is set to 1. If the [n] value is set to a value other than 1, the selected trace data results will be returned. See each command for details of what types of scalar results or trace data results are available. The binary data formats should be used for handling large blocks of data since they are smaller and transfer faster then the ASCII format. (FORMat:DATA)
FETCh may be used to return results other than those specified with the original READ or MEASure command that you sent.
INITiate Commands:
:INITiate:<measurement>
This command is not available for measurements in all the instrument modes:
For example, suppose you have previously initiated the ACP measurement, but now you are running the channel power measurement. If you send INIT:ACP? it will change from channel power to ACP and will initiate an ACP measurement.

MEASURE (Spectrum Analysis Mode)
READ Commands:
:READ:<measurement>[n]?
For example, suppose you have previously initiated the ACP measurement, but now you are running the channel power measurement. Then you send READ:ACP? It will change from channel power back to ACP and, using the previous ACP settings, will initiate the measurement and return results.
If the optional [n] value is not included, or is set to 1, the scalar measurement results will be returned. If the [n] value is set to a value other than 1, the selected trace data results will be returned. See each command for details of what types of scalar results or trace data results are available. The binary data formats should be used when handling large blocks of data since they are smaller and faster then the ASCII format. (FORMat:DATA)
Turns the measurement function off, so no measurement is currently active. When Meas Off is selected there are no functions under the Meas Setup key.
:CONFigure:SANalyzer
Remote Command Notes: :CONFigure:SANalyzer causes the present measurement to exit and places the analyzer in spectrum analysis mode.

MEASURE (Spectrum Analysis Mode)
Channel Power measures the power and power spectral density in the channel bandwidth that you specify. One marker pair on the display indicates the edges of the channel bandwidth. The center frequency, reference level, and channel bandwidth must be set by the user. For more information see “Meas Setup (Channel Power—CHP)” on page 399.
You can configure measurement settings by pressing Meas Setup after selecting Channel Power. You can make a measurement in single or continuous sweep mode. Pressing Meas Control enables you to pause or restart a measurement, or toggle between continuous and single measurement. To set Ref Level automatically, use Optimize Ref Level.
Figure 2 Channel Power Measurement Results

Pressing Meas Setup after Channel Power has been selected will access the channel power measurement setup menu. Pressing Radio Standard after Mode Setup has been selected will access all the Radio Standards available for which this measurement can be applied. Pressing Meas Control after Channel Power has been selected will access the channel power measurement control menu which allows you to pause or restart your measurement, or toggle between continuous and single measurement.
When a DVB-T radio standard is selected, this measurement will also compare the input signal against a pre-defined spectrum mask. The mask amplitudes defined in the DVB-T specification (ETSI EN 300 744 V1.4.1) range from 0 dB to 100 dB. This is for power measured in a 4kHz bandwidth therefore an adjustment to these values will need to be made as the measurement is made in a 7.61MHz bandwidth.
The measurement makes two sweeps, one in a 4kHz bandwidth and one in a 7.61MHz bandwidth, then computes the difference between the two. This value will be used to adjust the mask settings. As this value is calculated as part of the measurement setup, it is

MEASURE (Spectrum Analysis Mode)
expected that the DVB-T signal to be measured will be present prior to the standard being selected. The mask will be positioned relative to the power measured in the 7.61MHz bandwidth. The pass/fail indicator will be used to determine if the input signal exceeds the mask or not.

MEASURE (Spectrum Analysis Mode)
If the current span setting is narrower than the mask width, the mask will not be displayed and the message 10191 displayed. Increase the span to at least the mask width to display the mask. When the mask is not displayed, the pass/fail functionality will not be executed therefore no pass/fail flag will be displayed.
To ensure accuracy this measurement requires a specific minimum Sweep Point to Span ratio. This is calculated using; Sweep Points = Span / (Closest Mask Breakpoint / 2), then rounded off.
For DVB-T measurements using 24 MHz Span, the number of Sweep Points is 2001. If the Sweep Point to Span ratio is changed to a value less than the value obtained using the above algorithm, the message 10192 is then displayed, although the measurement is still made using the new sweep point to span ratio.
:CONFigure:CHPower
:INITiate:CHPower
:FETCh:CHPower? :MEASure:CHPower? :READ:CHPower?
Returns two values that correspond to the Channel Power and Power Spectral Density.
:FETCh:CHPower:CHPower? :MEASure:CHPower:CHPower? :READ:CHPower:CHPower?
Returns a single value that corresponds to the Channel Power.
:FETCh:CHPower:DENSity? :MEASure:CHPower:DENSity? :READ:CHPower:DENSity?
Returns a single value that corresponds to the Power Spectral Density.
Remote Command Notes: The main channel power is returned in the current amplitude units, and the density value is returned in current amplitude units/Hz
Example: FETC:CHP? or MEAS:CHP? or READ:CHP? command returns scalar results of main channel power and power density.
FETC:CHP:CHP? or MEAS:CHP:DENS? or READ:CHP:DENS? commands will return the single scalar result specified.
Measurement Results Available · Command Results Returned · Not Applicable

MEASURE (Spectrum Analysis Mode)
Occupied Bandwidth integrates the power of the displayed spectrum and puts markers at the frequencies between which a selected percentage of the power is contained. The measurement defaults to 99% of the occupied bandwidth power. The power-bandwidth routine first computes the combined power of all signal responses contained in the trace. For 99% occupied power bandwidth, markers are placed at the frequencies on either side of 99% of the power. 1% of the power is evenly distributed outside the markers. The difference between the marker frequencies is the 99% power bandwidth and is the value displayed. For more information see “Meas Setup (Occupied Bandwidth—OBW)” on page 451.
The occupied bandwidth function also indicates the difference between the analyzer center frequency and the center frequency of the channel, referred to as “Transmit Freq Error.” The measurement can be made in single or continuous sweep mode. The center frequency, reference level, and channel spacing must be set by the user.
To configure measurement settings, press Meas Setup after selecting Occupied BW. Pressing Radio Standard after Mode Setup has been selected will access all the Radio Standards available for which this measurement can be applied. Pressing Meas Control enables you to pause or restart a measurement, or toggle between continuous and single measurement.
:CONFigure:OBW
:INITiate:OBW
:FETCh:OBW? :MEASure:OBW? :READ:OBW?
Returns two values that correspond to the Occupied BW and Transmit Frequency Error in Hertz.
:FETCh:OBW:OBWidth? :MEASure:OBW:OBWidth? :READ:OBW:OBWidth?
Returns a single value that corresponds to the Occupied BW in Hertz.
:FETCh:OBW:FERRor?:MEASu re:OBW:FERRor? :READ:OBW:FERRor?
Returns a single value that corresponds to the Transmit Frequency Error in Hertz.
:FETCh:OBW:XDB? :MEASure:OBW:XDB? :READ:OBW:XDB?
Returns a single value that corresponds to the x dB BW in Hertz.
Remote Command Notes: The results for both values are returned in Hz.
Example: FETC:OBW? or MEAS:OBW? or READ:OBW? command returns scalar results of occupied bandwidth and transmit frequency error.
FETC:OBW:OBW? or MEAS:OBW:FERR? or READ:OBW:FERR? commands
Measurement Results Available · Command Results Returned · Not Applicable

MEASURE (Spectrum Analysis Mode)
will return the single scalar result specified.
Adjacent Channel Power (ACP) is a measure of the power that leaks into adjacent transmit channels. Depending on the radio standard selected from the Mode Setup menu, this measurement can run in several different modes in order to measure different types of signals. For more information see “Meas Setup (Adjacent Channel Power—ACP)” on page 351.
Turns on the adjacent channel power measurement. The center frequency, reference level, and channel bandwidth must be set by the user. The span is set according to the frequency and bandwidth of the offsets that are on, and whether the RRC filter is on. The screen is split and the lower window displays the absolute power in the main channel in dBm and the power in each of the adjacent channels in both dBm and dB relative to the main channel power. Also displayed for each offset are offset frequency and reference bandwidth.
The measurement settings may be configured by pressing Meas Setup after ACP has been selected. Pressing Radio Standard after Mode Setup has been selected will access all the Radio Standards available for which this measurement can be applied. Pressing Meas Control allows you to pause or restart your measurement, or toggle between continuous and single measurement mode.

MEASURE (Spectrum Analysis Mode)
Condition Command Results Returned
:CONFigure:ACPower
:INITiate:ACPower
Radio Std = None and
:FETCh:ACPower[n]? :MEASure:ACPower[n]? :READ:ACPower[n]?
Returns 3 comma-separated values that correspond to:
only offset A is on
main channel power, lower-adjacent channel power (dBc), and upper-adjacent channel power (dBc).
Radio Std ≠None or Radio Std = None and
Returns 28 comma-separated values that correspond in the following order:
more than one offset is on
Meas Type = Total Pwr Ref
Main channel relative power is relative to the main channel absolute power, and therefore is always equal to 0.00 dB.
Measurement Results Available · Not Applicable

MEASURE (Spectrum Analysis Mode)
Measurement Results Available (Continued)
Condition Command Results Returned
Radio Std ≠None or Radio Std = None and more than one offset is on
Returns 28 values that correspond in the following order:
Meas Type = Total Pwr Ref
Main channel relative power is relative to the main channel power spectral density, and therefore is always equal to 0.00 dB.
Remote Command Notes: The main channel power is returned in the current amplitude units, and the lower and upper channel results are always returned in dB.
Example: FETC:ACP? or MEAS:ACP? or READ:ACP? commands return the scalar results of main channel power, lower channel power (relative), and upper channel power (relative) if only one offset is set to on and the Radio Standard is set to None. Otherwise, 28 values are returned. If Meas Type is Total Power Ref, these are the main channel power in dBm and in dBc from the carrier (0 dBc) each repeated, followed by the absolute and relative power levels for each lower and upper offset if Meas Type is Total Power Ref. Otherwise they are the main channel power spectral density and relative power spectral density (0 dB) each repeated, followed by the absolute and relative spectral density values for each lower and upper offset.

MEASURE (Spectrum Analysis Mode)

MEASURE (Spectrum Analysis Mode)
Multi-Carrier Power is the measure of the power that leaks into adjacent transmit channels when two or more carriers are present. The results reported are identical to the adjacent power measurement, but the setup is different to allow for two or more carriers. For more information see “Meas Setup (Multi-Carrier Power—MCP)” on page 429.
The measurement settings may be configured by pressing Meas Setup after Multi-Carrier Power has been selected. Pressing Meas Control allows you to pause or restart your measurement, or toggle between continuous and single measurement mode.

MEASURE (Spectrum Analysis Mode)
Command n Results Returned
:CONFigure:MCPow er
:FETCh:MCP[n]? :MEASure:MCP[n]? :READ:MCP[n]?
n=1 (or not specified)
Returns 20 comma-separated scalar results, in the following order.
If the results are not available, −999.0 is returned.
:FETCh:MCPower? :MEASure:MCPower? :READ:MCPower?
n=2...13 Returns absolute and relative values for carrier n-1. If results are not available −999.0 is returned
Measurement Results Available · N/A Not Applicable

MEASURE (Spectrum Analysis Mode)
Measurement Results Available (Continued)
Command n Results Returned
:FETCh:MCPower? :MEASure:MCPower? :READ:MCPower?
n=14 Returns 36 comma-separated scalar results, in the following order.
1 to 24. All carriers absolute and relative values
If the results are not available, −999.0 is returned.
Example: FETC:MCP? or MEA:MCP? or READ:MCP?

MEASURE (Spectrum Analysis Mode)
The CCDF measurement is a statistical measurement of a high-level signal or peak power. It shows in both graphical and tabular form for what percentage of the time a signal exceeds its average power, and by how much this average is exceeded. For more information see “Meas Setup (Complementary Cumulative Distribution Function—CCDF)” on page 381.
For each single sweep of this measurement, the average power is defined by:
Pavg V2 Z0 ⁄ ( )
where V is the envelope voltage, Z0 is the characteristic impedance, and n is the number of samples (32k maximum).
Pavg is used as a guideline to define the probability of the samples, which is xdB above the average power. The number of samples collected per sweep depends upon the sampling rate and the setting of the measurement interval. Multiple sweeps may be required if the sample number exceeds the number of samples collected per sweep. The results and the CCDF curve are updated after each sweep.
Results from a 2.17 GHz 3GPP W-CDMA input signal are shown below in Figure 3.
Figure 3 Power Stat CCDF Measurement Results

n ⁄ =

MEASURE (Spectrum Analysis Mode)
Pressing Meas Setup after Power Stat CCDF has been selected will access the Power Stat CCDF measurement setup menu. The factory defaults provide a good starting point for this measurement. The settings are adjustable, however, to meet specific requirements. Pressing Radio Standard after Mode Setup has been selected will access all the Radio Standards available for which this measurement can be applied. Pressing Meas Control after Power Stat CCDF has been selected will access the Power Stat CCDF measurement control menu which allows you to pause or restart your measurement, or toggle between continuous and single measurement.

MEASURE (Spectrum Analysis Mode)
Command n Results Returned
:CONFigure:PSTatistic
:INITiate:PSTatistic
:FETCh:PSTatistic[n]? :MEASure:PSTatistic[n]? :READ:PSTatistic[n]?
n=1 (or not specified)
Returns 10 scalar results, in the following order.
n=2 Returns a series of 501 floating point numbers (in percent) that represent the current measured power statistics trace. This is the probability at specific power levels (average power), in the following order:
n=3 Returns a series of 501 floating point numbers (in percent) that represent the Gaussian trace. This is the probability at specific power levels (average power), in the following order:
Measurement Results Available · N/A Not Applicable

MEASURE (Spectrum Analysis Mode)
Measurement Results Available (Continued)
Command n Results Returned
n=4 Returns a series of 501 floating point numbers (in percent) that represent the user-definable reference trace. This is the probability that at specific power levels (average power) will occur, as follows:
Remote Command Notes: To save trace data use the fetch methods shown above.
Example: FETC:PST? or MEAS:PST? or READ:PST?
FETC:PST2? or MEAS:PST2? or READ:PST2?
FETC:PST3? or MEAS:PST3? or READ:PST3?
The range table state determines which harmonics are to be measured. If Range Table is set to Off, for each harmonic to be measured, the center frequency will be set to the appropriate integer multiple of the fundamental frequency and a sweep taken. The amplitude is calculated by averaging the power in the trace and is displayed in the results table in the results window. The desired Res BW value for each harmonic is calculated by multiplying the harmonic number with the Res BW of the fundamental. This results in the smallest Res BW available being used that is greater than or equal to the computed value. If a Res BW greater than the instrument maximum is calculated then the maximum Res BW of the instrument is used and an asterisk (*) is displayed after the amplitude result of the harmonic to indicate that the desired Res BW was not used. The span for each harmonic is set to zero.
If Range Table is set to On, for each range enabled after the fundamental, the center frequency will be set to the frequency of each range. If the current span is set to zero, the amplitude is calculated by averaging the power in the trace and is displayed in the results table in the results window. If the range has a non-zero span, the amplitude is calculated by making a band-power measurement over the span and is displayed in the results table in the results window.
After the specified number of harmonics have been measured and their amplitudes displayed, the total harmonic distortion is computed and displayed in the left-hand portion of the window (view is set to Harmonics & THD). If the measurement is in single mode, the display will remain at the last harmonic setting and the analyzer sweep will be stopped. If

MEASURE (Spectrum Analysis Mode)
the measurement is in continuous mode, the measurement will continually measure the amplitude of the fundamental and harmonics.
For averaged measurements, you may specify the number of sweeps over which the amplitude of each harmonic is measured. This means that if the Avg Number is set to 10, ten sweeps will be taken and averaged for each harmonic to be measured. The total harmonic distortion would then be calculated from these averaged amplitude readings. Note that averaging applies only to the amplitude of the harmonic, and does not affect the frequency at which the amplitude is measured.
Command n Results Returned
:CONFigure:HARMonics
:INITiate:HARMonics
:FETCh:HARMonics:AMPLitude:ALL?
N/A Returns the amplitude values of the first ten harmonics. The first value (for the fundamental) is measured in dBm. The remaining harmonics are measured in dBm from the fundamental. If fewer than ten harmonics are measured, zero is returned for any harmonic not measured.
:MEASure:HARMonics:AMPLitude:A LL?
:READ:HARMonics:AMPLitude:ALL?
:FETCh:HARMonics:AMPLitude[n]?
n=1 to 10 (n=1 for n not specified)
Returns the amplitude of the specified harmonic number n, where n=1 returns the amplitude of the fundamental in units of dBm.
:MEASure:HARMonics:AMPLitude[n] ?
:READ:HARMonics:AMPLitude[n]?
For example, n = 9 returns the amplitude of the ninth harmonic measured in dBc from the fundamental.
:FETCh:HARMonics[:DISTortion]?
N/A Returns the computed total harmonic distortion as a percentage.
:MEASure:HARMonics[:DISTortion]?
:READ:HARMonics[:DISTortion]?
:FETCh:HARMonics:FREQuency:ALL ?
N/A Returns the frequency values of the first ten harmonics, in Hz. The first harmonic is the fundamental. If fewer than ten harmonics are measured, zero is returned for any harmonic not measured.
:MEASure:HARMonics:FREQuency:A LL?
:READ:HARMonics:FREQuency:ALL?
Measurement Results Available · N/A Not Applicable

MEASURE (Spectrum Analysis Mode)
Command n Results Returned
:FETCh:HARMonics:FREQuency[n]?
n=2 to 10 (n=1 for n not specified)
Returns the frequency of the specified harmonic number N (in Hz).
:MEASure:HARMonics:FREQuency[n] ?
:READ:HARMonics:FREQuency[n]?
:FETCh:HARMonics:FUNDamental?
N/A Returns the frequency of the fundamental, measured in Hz.
:MEASure:HARMonics:FUNDamental ?
:READ:HARMonics:FUNDamental?
:FETCh:HARMonics[n]? :MEASure:HARMonics[n]? :READ:HARMonics[n]?
n=1 to 2 (n=1 for n not specified)
These commands return the power of the spurious emissions at the selected harmonic frequencies of the current channel frequency using the selected resolution bandwidth filter. You can set the number of harmonics to be measured. For each harmonic, the reference level is set 40 dB above the largest amplitude limit with input attenuation held constant. The attenuation is set to limit the carrier (fundamental) power at the spectrum analyzer input mixer to the value determined by Max Mixer Lvl.
not specified or n=1 Returns one value that corresponds to the total harmonic distortion (RMS voltage %).
n=2 Returns one value that corresponds to the total harmonic distortion (dB).
Example: FETC:HARM:AMPL:ALL?
MEAS:HARM:AMPL2 returns the amplitude of the second harmonic measured in dBc from the fundamental.
READ:HARM:FREQ10 returns the frequency of the tenth harmonic in Hz.
Measurement Results Available

MEASURE (Spectrum Analysis Mode)
The burst power measurement is an accurate method of determining the average power for the specified burst. The analyzer is set into zero-span mode, with a sweep time that captures at least one burst. For more information see “Meas Setup (Burst Power)” on page 371.

The burst power measurement acquires data from the “Measured Area” above when a radio standard is chosen and when Meas Setup, Meas Method, Measured Burst Width is selected. When Meas Setup, Meas Method, Above Threshold Lvl is selected, the “Measured Area” extends the burst width delimiter lines to the start and stop points.
The mean carrier power is calculated by:
Pavg ( ) 10 10
m n –

MEASURE (Spectrum Analysis Mode)
where Pavg = average power, n is the start trace point, m= the stop trace point, and p= the trace point amplitude power in dBm.
Figure 4 shows an example of the results obtained when measuring a Bluetooth™signal and with a user-defined burst width.
Figure 4 Burst Power Measurement Results

Pressing Meas Setup after Burst Power has been selected displays the burst power measurement setup menu. Pressing Meas Control after Burst Power has been selected displays the burst power control menu, where you can pause or restart a measurement, or toggle between continuous and single measurement.

MEASURE (Spectrum Analysis Mode)
Command n Results Returned
:CONFigure:BPOWer
:INITiate:BPOWer
:FETCh:BPOWer[n]? :MEASure:BPOWer[n]? :READ:BPOWer[n]?
n=1 (or not specified)
Returns 10 scalar results, in the following order.
n=2 Returns the displayed RF envelope trace data separated by commas. The number of data values is determined by the setting of the trace points parameter.
Example: FETC:BPOW? or MEAS:BPOW? or READ:BPOW?
Measurement Results Available · N/A Not Applicable

MEASURE (Spectrum Analysis Mode)
The third order intermodulation (TOI) measurement computes and displays the output intercept point (IP3), and places markers upon the trace to indicate the measured signals and third-order products. For more information see “Meas Setup (Intermod (TOI))” on page 423.
:CONFigure:TOIN
:INITiate:TOIN
:FETCh:TOIN? :MEASure:TOIN? :READ:TOIN?
Returns 6 scalar results, in the following order.
:FETCh:TOIN:IP2? :MEASure:TOIN:IP2? :READ:TOIN:IP2?
Returns 12 scalar results, in the following order.
1. The worst case Output Intermod Point value in Hz. 2. The worst case Output Intermod Power value in dBm. 3. The worst case Output Intercept Power value in dBm 4. The lower base frequency value in Hz 5. The lower base power value in dBm 6. The upper base frequency value in Hz 7. The upper base power value in dBm 8. The lower Output Intermod Point in Hz 9. The lower Output Intermod Power value in dBm 10. The lower Output Intercept Power value in dBm 11. The upper Output Intermod Power value in dBm 12. The upper Output Intermod Point in Hz 13. The upper Output Intercept Power value in dBm
:FETCh:TOIN:IP3? :MEASure:TOIN:IP3? :READ:TOIN:IP3?
Returns the worst case Output Intercept Power value in dBm.
Example: FETC:TOIN? or MEAS:TOIN? or READ:TOIN?
FETC:TOIN:IP3? or MEAS:TOIN:IP3? or READ:TOIN:IP3?
Measurement Results Available · Command Results Returned · Not Applicable

MEASURE (Spectrum Analysis Mode)
The spurious emissions measurement identifies and determines the power level of spurious emissions in certain frequency bands. For more information see “Meas Setup (Spurious Emissions)” on page 485.
:CONFigure:SPURious
:INITiate:SPURious
:FETCh:SPURious[n]?
:MEASure:SPURious[n]?
:READ:SPURious[n]?
n=1 (or not specified)
Returns a variable-length list of values containing detailed spur information. The total number of data values returned is (1 + 6x spurs), where x is the number of spurs identified. You can have up to 121 data entries if you have the maximum 20 spurs identified. The values returned are in the following order:
[Repeat the following for each spur]
n=2 through 21 Returns trace data values for the selected frequency range (where range number = n −1). If the selected range is not active, then 9.91E37 is returned for each non-active trace data element.
n=22 Returns the number of spurs found.
Example: MEAS:SPUR?
Spectrum Emission Mask measurement includes the in-band and out-of-band spurious emissions. As it applies to W-CDMA (3GPP), this is the power contained in a specified frequency bandwidth at certain offsets relative to the total carrier power. It may also be expressed as a ratio of power spectral densities between the carrier and the specified offset frequency band. For WLAN, the reference power is taken as the peak PSD in the signal,
n Results Returned

MEASURE (Spectrum Analysis Mode)
and all offset results are also the peak PSD in that offset as opposed to the integrated power.
Offsets that are turned off (inactive) will return -999.0 when their results are queried over SCPI.
For more information see “Meas Setup (Spectrum Emissions Mask—SEM)” on page 457.

MEASURE (Spectrum Analysis Mode)
:CONFigure:SEMask
:INITiate:SEMask
:FETCh:SEMask[n]? :MEASure:SEMask[n]? :READ:SEMask[n]?
n=1 (or not specified)
Returns 60 scalar results, in the following order:
1. Reserved for the future use, returns –999.0 2. Absolute power at the center frequency (reference) area (dBm) 3. Reserved for the future use, returns –999.0 4. Reserved for the future use, returns –999.0 5. Peak frequency in the center frequency (reference) area (Hz) 6. Reserved for the future use, returns –999.0 7. Reserved for the future use, returns –999.0 8. Reserved for the future use, returns –999.0 9. Reserved for the future use, returns –999.0 10.Reserved for the future use, returns -999.0. 11.Relative power on the negative offset A (dBc) 12.Absolute power on the negative offset A (dBm) 13.Relative peak power on the negative offset A (dBc) 14.Absolute peak power on the negative offset A (dBm) 15.Peak frequency in the negative offset A (Hz) 16.Relative power on the positive offset A (dBc) 17.Absolute power on the positive offset A (dBm) 18.Relative peak power on the positive offset A (dBc) 19.Absolute peak power on the positive offset A (dBm) 20.Peak frequency in the positive offset A (Hz) 21.Relative power on the negative offset B (dBc) … 59.Absolute peak power on the positive offset E (dBm) 60.Peak frequency in the positive offset E (Hz)
Using Total Power Reference

MEASURE (Spectrum Analysis Mode)
Measurement Results Available (Continued)
Using Power Spectral Density Reference
Returns 60 scalar results, in the following order:
n=1 (or not specified)
1. Reserved for the future use, returns –999.0 2. Absolute power at the center frequency (reference) area (dBm) 3. Reserved for the future use, returns –999.0 4. Reserved for the future use, returns –999.0 5. Peak frequency in the center frequency (reference) area (Hz) 6. Reserved for the future use, returns –999.0 7. Reserved for the future use, returns –999.0 8. Reserved for the future use, returns –999.0 9. Reserved for the future use, returns –999.0 10.Reserved for the future use, returns -999.0. 11.Relative power on the negative offset A (dB) 12.Absolute power on the negative offset A (dBm/Hz) 13.Relative peak power on the negative offset A (dB) 14.Absolute peak power on the negative offset A (dBm/Hz) 15.Peak frequency in the negative offset A (Hz) 16.Relative power on the positive offset A (dB) 17.Absolute power on the positive offset A (dBm/Hz) 18.Relative peak power on the positive offset A (dB) 19.Absolute peak power on the positive offset A (dBm/Hz) 20.Peak frequency in the positive offset A (Hz) 21.Relative power on the negative offset B (dB) … 59.Absolute peak power on the positive offset E (dBm/Hz) 60.Peak frequency in the positive offset E (Hz)
n=2 Returns the displayed frequency domain spectrum trace data separated by comma. The number of data points is determined by the setting of the trace points parameter.
n=3 Returns the displayed frequency domain absolute limit trace data separated by comma. The number of data points is determined by the setting of the trace points parameter.
n=4 Returns the displayed frequency domain relative limit trace data separated by comma. The number of data points is determined by the setting of the trace points parameter.
Command / Condition · n Results Returned · (Continued)

MEASURE (Spectrum Analysis Mode)
Measurement Results Available (Continued)
Using Total Power Reference
n=5 Total Power Reference
Returns 12 scalar values (in dBm) of the absolute power of the segment frequencies:
Using Power Spectral Density Reference
n=5 Power Spectral Density Reference
Returns 12 scalar values (in dBm/Hz) of the absolute power of the segment frequencies:
Using Total Power Reference
n=6 Total Power Reference
Returns 12 scalar values (in dBc) of the power relative to the carrier at the segment frequencies:
Using Power Spectral Density Reference
n=6 Power Spectral Density Reference
Returns 12 scalar values (in dBc/Hz) of the power relative to the carrier at the segment frequencies:
Command / Condition · n Results Returned

MEASURE (Spectrum Analysis Mode)
Measurement Results Available (Continued)
n=7 Returns 12 pass/fail test results (0 = passed, or 1 = failed) determined by testing the absolute power in each offset against the specified offset’s absolute power limits:
n=8 Returns 12 scalar values of the pass/fail (0 = passed, or 1 = failed) results determined by testing the relative power in each offset against the specified offset’s relative power limits:
n=9 Returns 12 scalar values of frequency (in Hz) that have peak power in each offset:
n=10 Returns 12 scalar values (in dBm) of the absolute peak power of the segment frequencies:
Command / Condition · n Results Returned

MEASURE (Spectrum Analysis Mode)
Measurement Results Available (Continued)
n=11 Returns 12 scalar values (in dBc) of the peak power relative to the carrier at the segment frequencies:
n=12 Returns the band power result when in WLAN standard. If not in WLAN standard the value returned will be .999.0
Example: FETC:SEM? or MEA:SEM? or READ:SEM?
This command returns the name of the measurement that is currently running.
:CONFigure?
Example: CONF?
Command / Condition · n Results Returned

These functions allow you to pause and resume the currently selected measurement and to select between continuous or single measurements.
This function restarts a previously paused measurement at the beginning. If the current measurement is still in process, it will stop it as soon as possible and restart it from the beginning.
Key Path: Front-panel key. It can also be found under Meas Control.
:INITiate:RESTart
Remote Command Notes: This command is equivalent to sending an :ABORt command followed by an :INITiate[:IMMediate] command. See “Abort the Sweep or Measurement (Remote Command Only)” on page 331. for more information.
Switches the analyzer between triggering the current measurement/sweep continuously or triggering a single measurement. The front panel Single key also puts the analyzer in single-measurement mode.
Meas Control

Use :INITiate:CONTinuous OFF|ON See “SWEEP” on page 241.
Remote Command Notes: This command affects sweeping when in the SA mode. It affects measurements when a measurement has been selected from the MEASure command subsystem.
This function pauses the currently running measurement. Pressing Pause will toggle between pausing and resuming your measurement. The key label will toggle between Pause and Resume. If an averaged measurement was in progress, the average counter is frozen when the measurement is halted
:INITiate:PAUSe to pause the measurement
:INITiate:RESume to resume the measurement.
Remote Command Notes: See “Abort the Sweep or Measurement (Remote Command Only)” on page 331. for more information.
This command initiates a sweep if in SA mode with no measurement currently selected.
The command is ignored if the instrument is in a measurement (selected under the MEASURE key), but the measurement is currently running, (INITiate:CONTinuous ON).
If a measurement is selected but it is in the idle state (i.e. it’s not running, INITiate:CONT OFF), this command triggers the instrument, when trigger conditions are met. The trigger system is initiated, it completes one full trigger cycle and returns to the “waiting” state. Depending on the measurement selected and the number of averages, there may be
Meas Control

multiple data acquisitions, with multiple trigger events, for one full trigger cycle. The instrument must have external triggering selected, or the command will be ignored. Use the TRIGer[:SEQuence]:SOURce EXT command to select the external trigger.
:INITiate[:IMMediate]
Remote Command Notes: See also the *TRG command and the TRIGger subsystem.
Use the [:SENSe]::TRIGger:SOURce command to select the desired trigger. The instrument must be in the single measurement mode. If :INITiate:CONTinuous is ON then the command is ignored.
Use :FETCh? to transfer a measurement result from memory to the output buffer. Refer to individual commands in the MEASure subsystem for more information.
Stops any sweep or measurement in progress and resets the sweep or trigger system. A measurement refers to any of the measurements found in the MEASURE menu. If the trigger conditions are met, another sweep is initiated immediately.
If :INITiate:CONTinuous is off (single measure), then :INITiate:IMMediate will start a new single measurement.
If :INITiate:CONTinuous is on (continuous measure), a new continuous measurement begins immediately.
The INITiate and/or TRIGger subsystems contain additional related commands.
:ABORt
Remote Command Notes: In the continuous measurement mode, the Restart key is equivalent to ABORt.
Meas Control

Meas Control

Mode Setup (Spectrum Analysis Mode)
Enables you to change measurement settings common to all measurements in the MEASURE menu. In Spectrum Analysis mode, there are several built-in power measurements. Parameters that you set in the Mode Setup menu affect all of these measurements.
Accesses the radio standards key menu to enable you to select a radio standard. Selecting a radio standard modifies spectrum analyzer settings for the measurement activated under the MEASURE menu.
Added with PSA firmware revision A.02.00
[:SENSe]:RADio:STANdard[:SELect] NONE|IS95a|JSTD|IS95C|GSM|W3GPP|C2000MC1|C2000DS|C2000MC3|NADC|PDC|BLUEtoo th|TETRa |WL802DOT11A|WL802DOT11B|WL802DOT11G|HIPERLAN2||DVBTGPN|DVBTIPN
[:SENSe]:RADio:STANdard[:SELect]?
5.3.1.1 None
Selects no radio standard. If Radio Std, None is selected when a measurement is running, all measurement parameters are returned to their default values. Analyzer parameters outside the MEASURE or Meas Setup key menus are not affected.

Mode Setup (Spectrum Analysis Mode)
Use[:SENSe]:RADio:STANdard[:SELect] See “Radio Std” on page 333.
Sets the specific parameters for the selected measurement (located under the “MEASURE (Spectrum Analysis Mode)” key description) appropriate for industry standard IS-95A. The Spectrum Emissions Mask, Harmonic Distortion, and Spurious Emissions measurements are not available when this radio standard is selected.
Use[:SENSe]:RADio:STANdard[:SELect] See “Radio Std” on page 333.
Sets the specific parameters for the selected measurement (located under the “MEASURE (Spectrum Analysis Mode)” key description) appropriate for industry standard J-STD-008. The Spectrum Emissions Mask, Harmonic Distortion, MCP, and Spurious Emissions measurements are not available when this radio standard is selected.
Use[:SENSe]:RADio:STANdard[:SELect] See “Radio Std” on page 333.
Sets the specific parameters for the selected measurement (located under the “MEASURE (Spectrum Analysis Mode)” key description) appropriate for industry standard IS-95C. The Spectrum Emissions Mask, Harmonic Distortion, and Spurious Emissions measurements are not available when this radio standard is selected.

Mode Setup (Spectrum Analysis Mode)
Added with PSA firmware revision A.03.00
Use[:SENSe]:RADio:STANdard[:SELect] See “Radio Std” on page 333.
Sets the specific parameters for the selected measurement (located under the “MEASURE (Spectrum Analysis Mode)” key description) appropriate for industry standard GSM/EDGE. Power Stat CCDF and Burst Power are the only measurements available for this standard.
Use[:SENSe]:RADio:STANdard[:SELect] See “Radio Std” on page 333.
Sets the specific parameters for the selected measurement (located under the “MEASURE (Spectrum Analysis Mode)” key description) appropriate for industry standard W-CDMA. All measurements except Harmonic Distortion and Spurious Emissions are available for this standard.
Use[:SENSe]:RADio:STANdard[:SELect] See “Radio Std” on page 333.
5.3.1.7 cdma2000 MC-1X
Sets the specific parameters for the selected measurement (located under the “MEASURE (Spectrum Analysis Mode)” key description) appropriate for industry standard cdma2000-MC-1X. All measurements except Harmonic Distortion, Spurious Emissions,

Mode Setup (Spectrum Analysis Mode)
and Spectrum Emission Mask are available for this standard.
Added with PSA firmware revision A.02.00
Use[:SENSe]:RADio:STANdard[:SELect] See “Radio Std” on page 333.
5.3.1.8 cdma2000 MC-3X
Sets the specific parameters for the selected measurement (located under the “MEASURE (Spectrum Analysis Mode)” key description) appropriate for industry standard cdma2000:MC-3X. All measurements except Multi-Carrier Power, Harmonic Distortion, Spurious Emissions and Spectrum Emission Mask are available for this standard.
Added with PSA firmware revision A.02.00
Use[:SENSe]:RADio:STANdard[:SELect] See “Radio Std” on page 333.
5.3.1.9 cdma2000 DS
Sets the specific parameters for the selected measurement (located under the “MEASURE (Spectrum Analysis Mode)” key description) appropriate for industry standard cdma2000:DS. All measurements except Multi-Carrier Power, Harmonic Distortion, Spurious Emissions and Spectrum Emission Mask are available for this standard.
Added with PSA firmware revision A.02.00
Use[:SENSe]:RADio:STANdard[:SELect] See “Radio Std” on page 333.

Mode Setup (Spectrum Analysis Mode)
Sets the specific parameters for the selected measurement (located under the “MEASURE (Spectrum Analysis Mode)” key description) appropriate for industry standard NADC. Channel Power and Intermod TOI measurements are not available when Device (MS) is selected. All other measurements are available when this radio standard is selected except: Spectrum Emissions Mask, Harmonic Distortion, Spurious Emissions, and Multi-Carrier Power.
Added with PSA firmware revision A.02.00
Use[:SENSe]:RADio:STANdard[:SELect] See “Radio Std” on page 333.
Sets the specific parameters for the selected measurement (located under the “MEASURE (Spectrum Analysis Mode)” key description) appropriate for industry standard PDC. The Burst Power measurement is not available when Device (BTS) is selected. The Channel Power and Intermod TOI measurements are not available when Device (MS) is selected. All other measurements except Multi-Carrier Power, Harmonic Distortion, Spurious Emissions and Spectrum Emission Mask measurements are available for this standard.
Use[:SENSe]:RADio:STANdard[:SELect] See “Radio Std” on page 333.
5.3.1.12 Bluetooth™
Sets the specific parameters for the selected measurement (located under the “MEASURE (Spectrum Analysis Mode)” key description) appropriate for industry standard Bluetooth™.

Mode Setup (Spectrum Analysis Mode)
Power Stat CCDF and Burst Power are the only measurements available for this standard.
Use[:SENSe]:RADio:STANdard[:SELect] See “Radio Std” on page 333.
Sets the specific parameters for the selected measurement (located under the “MEASURE (Spectrum Analysis Mode)” key description) appropriate for industry standard TETRA. This standard is only available for the Channel Power and ACP measurements.
Use[:SENSe]:RADio:STANdard[:SELect] See “Radio Std” on page 333.
Accesses the W-LAN radio standards key menu to enable you to select a W-LAN standard. Selecting a radio standard modifies spectrum analyzer settings for the CHP or CCDF measurement activated under the MEASURE menu.
5.3.1.14.1 802.11a Sets the specific parameters for the selected measurement (located under the “MEASURE (Spectrum Analysis Mode)” key description) appropriate for industry standard 802.11a. This standard is only available for the Spectrum Emission Mask (SEM) measurement.
Added with PSA firmware revision A.03.00
Use[:SENSe]:RADio:STANdard[:SELect] See “Radio Std” on page 333.

Mode Setup (Spectrum Analysis Mode)
5.3.1.14.2 802.11b Sets the specific parameters for the selected measurement (located under the “MEASURE (Spectrum Analysis Mode)” key description) appropriate for industry standard 802.11b. This standard is only available for the Spectrum Emission Mask (SEM) measurement.
Added with PSA firmware revision A.03.00
Use[:SENSe]:RADio:STANdard[:SELect] See “Radio Std” on page 333.
5.3.1.14.3 802.11g Sets the specific parameters for the selected measurement (located under the “MEASURE (Spectrum Analysis Mode)” key description) appropriate for industry standard 802.11g. This standard is only available for the Spectrum Emission Mask (SEM) measurement.
Use[:SENSe]:RADio:STANdard[:SELect] See “Radio Std” on page 333.
5.3.1.14.4 Hiper LAN/2 Sets the specific parameters for the selected measurement (located under the “MEASURE (Spectrum Analysis Mode)” key description) appropriate for industry standard Hiper LAN2. This standard is only available for the Spectrum Emission Mask (SEM) measurement.
Use[:SENSe]:RADio:STANdard[:SELect] See “Radio Std” on page 333.

Mode Setup (Spectrum Analysis Mode)
Accesses the DVB-T key menu to enable you to select a DVB-T mask filtering standard. Selecting a DVB-T filtering standard modifies spectrum analyzer settings for the CHP or CCDF measurement activated under the MEASURE menu.
5.3.1.15.1 L/SECAM/NICAM Pressing L/SECAM/NICAM selects the filtering masks for the L/SECAM/NICAM analog television system.
5.3.1.15.2 G/PAL/NICAM Pressing G/PAL/NICAM selects the filtering masks for the G/PAL/NICAM analog television system.
5.3.1.15.3 I/PAL/NICAM Pressing I/PAL/NICAM selects the filtering masks for the I/PAL/NICAM analog television system.

Mode Setup (Spectrum Analysis Mode)
Accesses the key menu for selecting the device, packet type, or signal bandwidth to be measured.
There is no equivalent remote command.
5.3.2.1 Signal BW
Allows you to set the measurement bandwidth when Radio Std (None) is selected. This function is available only when you select None as the standard.
[:SENSe]:RADio:STANdard:SBWidth
[:SENSe]:RADio:STANdard:SBWidth?
5.3.2.2 Device BTS/MS
This function is only available when you have selected the standard: IS-95, J-STD-008, cdma2000, W-CDMA, NADC, PDC, TETRA, or GSM. It enables you to select either the base transmitter station (BTS) setup defaults or the mobile station (MS) defaults for the standard that you have selected. This key is unavailable when you select Bluetooth™.
[:SENSe]:RADio:STANdard:DEVice BTS|MS
[:SENSe]:RADio:STANdard:DEVice?

Mode Setup (Spectrum Analysis Mode)
5.3.2.3 Packet Type
This function is only available when the standard you have selected is Bluetooth™. It enables you to set the instrument settings for testing DH1, DH3, or DH5 packet type.
[:SENSe]:RADio:STANdard:PACKet DH1|DH3|DH5
[:SENSe]:RADio:STANdard:PACKet?
5.3.2.4 Band Class
This function is only available when you have selected the standard: IS-95C. It enables you to select the band class.
[:SENSe]:RADio:STANdard:BAND:CLASs BC0|BC1
[:SENSe]:RADio:STANdard:BAND:CLASs?
Enables you to retain the base-instrument parameter settings used by individual measurements that are typically lost when switching between measurements.
When Retain Params is set to Off, the values of the fully configurable parameters saved and recalled during a measurement change.
When Retain Params is set to On, the fully configurable parameters will be saved and recalled during a measurement change, along with Span, Ref Level, Single/Cont, Detector, Avg/VBW Type, Points, and if applicable, Spectrum/Bar/Combined and Graticule On/Off.

Mode Setup (Spectrum Analysis Mode)

Mode Setup (Spectrum Analysis Mode)
[:SENSe]:RADio:STANdard:RPARams[:STATe]ON|Off|1|0
[:SENSe]:RADio:STANdard:RPARams[:STATe]?
Allows you to specify whether all measurements and radio standards are enabled (set to Yes) or not (Set to No).
[:SENSe]:RADio:STANdard:EAMeas YES|NO
[:SENSe]:RADio:STANdard:EAMeas?
This command is the remote version of the Optimize Ref Level key. If this function is applicable to the currently selected measurement, this command sets the reference level and attenuator to optimum values based on the signal present at the input.

Mode Setup (Spectrum Analysis Mode)
[:SENSe]:POWer[:RF]:RANGe:AUTO ONCE

Mode Setup (Spectrum Analysis Mode)

This function restarts a previously paused measurement at the beginning. If the measurement is active, it will stop it as soon as possible and restart it from the beginning.If no measurement is active and Sweep (Single) is selected, a new sweep is initiated.
When in Average (On) mode, (BW/Avg, Average) the averaging function is restarted (the trace is reset and the average number is reset to zero).
Key Path: Front-panel key. It can also be found under Meas Control.
:INITiate:RESTart
Remote Command Notes: This command is equivalent to sending an :ABORt command followed by an :INITiate[:IMMediate] command. See “Abort the Sweep or Measurement (Remote Command Only)” on page 331. for more information.
Restart

Restart

If the analyzer is in continuous sweep mode and not in a measurement (Measure, Meas Off), pressing Single changes the sweep control to single sweep, and executes a sweep after the trigger condition is met. If the analyzer is already in single sweep, pressing Single executes a new sweep after the trigger condition is met.
Some instrument settings require more than one sweep to complete the measurement (see BW/Avg, Average), or if you have selected a measurement from the functions under the MEASURE key, this function sets the trigger system to be initiated only once. In this case the trigger condition can be met only once and then all the necessary sweeps will be executed to make the measurement or complete the averaging function.
With Average on (BW/Avg, Average (On)), pressing Single resets the average trace and starts the average again from a count of zero. Sweeps are averaged until N sweeps are then taken (where N is the average number), and then the sweep is halted.
:INITiate[:IMMediate]
Remote Command Notes: Use the :TRIGger[:SEQuence]:SOURce command to select the trigger source.
See also the Single/Continuous sweep function in “Sweep” on page 242 and the command INITiate:CONTinuous ON|OFF.
See also the *TRG command information in the STATus section.
Single

Single

Meas Setup (Adjacent Channel Power—ACP)
If the adjacent channel power (ACP) measurement has been selected in the Measure menu of the spectrum analysis mode, this key displays the ACP measurement setup menu.
The adjacent channel power measurement measures the power that leaks into adjacent transmit channels. Depending on the radio standard chosen from the Mode Setup menu, this measurement can run in several different modes in order to measure different types of signals.
History: Added with firmware revision A.02.00.
There is no equivalent remote command.
Pressing Avg Number to On enables you to specify the number of measurements that will be averaged when calculating the measurement result. The average will be displayed at the end of each sweep.Setting Avg Number to Off disables the measurement averaging.
State Saved: Saved in instrument state.
History: Added with firmware revision A.02.00.
[:SENSe]:ACPower:AVERage:COUNt
[:SENSe]:ACPower:AVERage:COUNt?
[:SENSe]:ACPower:AVERage[:STATe] OFF|ON|0|1 turns the averaging on or off.
[:SENSe]:ACPower:AVERage[:STATe]?
ACP:AVER OFF
ACP:AVER?

Meas Setup (Adjacent Channel Power—ACP)

Meas Setup (Adjacent Channel Power—ACP)
Press Avg Mode to select the type of termination control used for the averaging function as either Exp or Repeat. This determines the averaging action after the specified number of measurements (average count) is reached.
State Saved: Saved in instrument state.
History: Added with firmware revision A.02.00.
[:SENSe]:ACPower:AVERage:TCONrol EXPonential|REPeat
[:SENSe]:ACPower:AVERage:TCONrol?
5.6.3 Chan Integ BW
Press Chan Integ BW to specify the range of integration used in calculating the power in the main channel.
Key Notes: If RRC Filter is on, the actual integration bandwidth used is the displayed integration bandwidth multiplied by (1 + Filter Alpha).
State Saved: Saved in instrument state.
Factory Preset: 2.0 MHz or as defined by the selected radio standard.
Knob Increment: Span/50 if non-zero span.
RBW/100 if zero span.
Step Key Increment: RBW if CF Step is set to Auto with zero span.

Meas Setup (Adjacent Channel Power—ACP)
Span/10 if CF Step is set to Auto with non-zero span.
CF Step if CF Step is set to Manual.
Range: Depends on the adjacent-channel settings and the minimum and maximum analyzer span.
History: Added with firmware revision A.02.00.
[:SENSe]:ACPower:BANDwidth|BWIDth:INTegration
[:SENSe]:ACPower:BANDwidth|BWIDth:INTegration?
5.6.4 Offset/Limits
Accesses menu keys that allow you to configure the offsets for the ACP measurement.
5.6.4.1 Offset
Selects the offset the menu keys will affect. Press Offset until the letter of the desired offset is underlined.
State Saved: Saved in instrument state.
History: Added with firmware revision A.02.00.
There is no remote command for this function.
5.6.4.2 Offset Freq or Edge Freq
Enables you to set the frequency difference from the center of the main channel to the center of the offset for a maximum of 6 offsets (labeled A-F). It also allows you to turn on/off the offsets that you want to measure.

Meas Setup (Adjacent Channel Power—ACP)
State Saved: Saved in instrument state.
The following table shows the default array if no radio standard (None) is selected.

Meas Setup (Adjacent Channel Power—ACP)
Knob Increment: Span/50 if non-zero span.
RBW/100 if zero span.
Step Key Increment: RBW if CF Step is set to Auto with zero span.
Span/10 if CF Step is set to Auto with non-zero span.
CF Step if CF Step is set to Manual.
History: Added with firmware revision A.02.00.
[:SENSe]:ACPower:OFFSet:LIST[:FREQuency] ,,,,,
[:SENSe]:ACPower:OFFSet:LIST[:FREQuency]?
[:SENSe]:ACPower:OFFSet:LIST:STATE OFF|ON|0|1, OFF|ON|0|1
[:SENSe]:ACPower:OFFSet:LIST:STATE?
Remote Command Notes: This command, along with commands [:SENSe]:ACPower:OFFSet:LIST:BANDwidth|BWIDth[:INTegration] and [:SENSe]:ACPower:OFFSet:LIST:STATe, are used to set the parameters for all of the offsets, up to six. The following table shows the values of these parameters if no radio standard (None) is selected.
Offset State Frequency Integ. BW
A ON 3 MHz 2 MHz
B OFF 0 2 MHz
C OFF 0 2 MHz
D OFF 0 2 MHz
E OFF 0 2 MHz
F OFF 0 2 MHz

Meas Setup (Adjacent Channel Power—ACP)
Example: Sending fewer than six parameters to one of these commands will leave the values of the unspecified offsets unchanged. If you don’t send settings for all 6 offsets, it will set all the offsets that you specified, then it will set any remaining offsets to the same setting as the last offset that you sent.
ACP:OFFS:LIST 50 Hz,75 Hz,100 Hz,125 Hz,150 Hz,175 Hz
For example, after the above command is sent, sending the command with only four parameters (ACP:OFFS:LIST:STAT ON,ON,ON,OFF) will result in the fifth and sixth offset remaining the same as the previous setting (OFF).
5.6.4.3 Ref BW
Sets the reference bandwidth (integration bandwidth) for each offset.
State Saved: Saved in instrument state.
Factory Preset: 2 MHz, if no radio standard (None) is selected.
History: Added with firmware revision A.02.00.
[:SENSe]:ACPower:OFFSet:LIST:BANDwidth|BWIDth[:INTegration] ,,,,,
[:SENSe]:ACPower:OFFSet:LIST:BANDwidth|BWIDth[:INTegration]?
5.6.4.4 Pos Offset Limit
Enables you to set the upper limit for the upper segment of the specified offset pair.
State Saved: Saved in instrument state.

Meas Setup (Adjacent Channel Power—ACP)
History: Added with firmware revision A.02.00.

Meas Setup (Adjacent Channel Power—ACP)
:CALCulate:ACPower:OFFSet:LIST:LIMit:POSitive[:UPPer]:DATA ,,,,,
:CALCulate:ACPower:OFFSet:LIST:LIMit:POSitive[:UPPer]:DATA?
5.6.4.5 Neg Offset Limit
Enables you to set the upper limit for the lower segment of the specified offset pair.
State Saved: Saved in instrument state.
History: Added with firmware revision A.02.00.
:CALCulate:ACPower:OFFSet:LIST:LIMit:NEGative[:UPPer]:DATA ,,,,,
:CALCulate:ACPower:OFFSet:LIST:LIMit:NEGative[:UPPer]:DATA?
5.6.5 Meas Type
Press Meas Type to specify the reference for the measurement, either Total Pwr Ref or PSD Ref. Relative values can be displayed referenced to either the total power (Total Pwr Ref) or the power spectral density (PSD Ref) measured in the main channel.
State Saved: Saved in instrument state.
History: Added with firmware revision A.02.00.
[:SENSe]:ACPower:TYPE PSDRef|TPRef

Meas Setup (Adjacent Channel Power—ACP)
[:SENSe]:ACPower:TYPE?
5.6.6 Optimize Ref Level
Sets the input attenuator to optimize the robustness of the measurement, in which it is free from errors due to input compression. This setting will not necessarily give the optimum dynamic range, nor the optimum accuracy. No single setting can optimize both, and the optimum setting often depends on the signal characteristics.
State Saved: Not saved.
History: Added with firmware revision A.02.00.
[:SENSe]:POWer[:RF]:RANGe:AUTO ONCE
Enables you to set the measurement method to either the integration bandwidth method (IBW) or the resolution bandwidth method (RBW). The resolution bandwidth method is most useful for measuring cdmaOne and cdma2000 signals; the integration bandwidth method is preferred for other signals.
State Saved: Saved in instrument state.
History: Added with firmware revision A.02.00.
[:SENSe]:ACPower:METHod IBW|RBW
[:SENSe]:ACPower:METHod?

Meas Setup (Adjacent Channel Power—ACP)

Meas Setup (Adjacent Channel Power—ACP)
Enables you to set the adjacent channel power reference to automatic or manual. When set to automatic, the carrier power result reflects the measured power value in the carrier. When set to manual, the last measured value is captured and held, or may be entered by the user. Relative values are displayed, referenced to the total power measured in the main channel.
State Saved: Saved in instrument state.
History: Added with firmware revision A.02.00.
[:SENSe]:ACPower:CARRier[:POWer]
[:SENSe]:ACPower:CARRier:AUTO[:STATe] OFF|ON|0|1
[:SENSe]:ACPower:CARRier:AUTO[:STATe]?
5.6.9 PSD Ref
Enables you to set the power spectral density in the carrier (main channel) that will be used to compute the relative power spectral density values for the offsets. When the PSD Ref state is set to Auto, this will be set to the measured carrier power spectral density.
Step Key Increment: 1, 1.5, 2...
History: Added with firmware revision A.02.00.
[:SENSe]:ACPower:CARRier:CPSD
[:SENSe]:ACPower:CARRier:CPSD?
Remote Command Notes: This function is only available when measurement type is set to PSD Ref, use the command [:SENSe]:ACPower:TYPE PSDRef to select the measurement type.

Meas Setup (Adjacent Channel Power—ACP)
Pressing Limit Test turns the testing of the limit line on or off. When Limit Test is set to On, each offset is compared to its upper and lower offset limit. In those cases where the power exceeds the limit, a red “F” is placed next to the dBc result to indicate a failure. If there are any failures, the pass/fail indicator area show “FAIL” in red; if there are none, it shows “PASS” in green. Any offsets that are in the off state (see “Offset Freq or Edge Freq” on page 354) are not measured and their results will not be displayed on screen.
State Saved: Saved in instrument state.
History: Updated with firmware revision A.03.00.
Added with firmware revision A.02.00.
[:SENSe]:ACPower:LIMit[:STATe] OFF|ON|0|1
[:SENSe]:ACPower:LIMit[:STATe]?
5.6.11 RRC Filter
Pressing Filter turns the Root Raised Cosine filter on or off. This filter is the type specified in the NADC and 3GPP W-CDMA standards and is only available when 3GPP W-CDMA or NADC is selected as the Radio Std from the Mode Setup menu. If 3GPP W-CDMA is selected, the roll-off value (alpha) for the filter will be initially set to 0.22. T is set to 1/RefBW (default 260 ns). If NADC is selected, the roll-off will be initially set to 0.35. T is set to 1/RefBW (default 42 µs). If TETRA has been selected, the roll-off will be initially set to 0.35. T is set to 1/RefBW (default 55.6 µs). The roll-off value can be changed using Filter Alpha.
The filter is unavailable when the measurement method is set to RBW.
State Saved: Saved in instrument state.
History: Added with firmware revision A.03.00.

Meas Setup (Adjacent Channel Power—ACP)
[:SENSe]:ACPower:FILTer[:RRC][:STATe] OFF|ON|0|1
[:SENSe]:ACPower:FILTer[:RRC][:STATe]?
5.6.12 Filter Alpha
Press Filter to input the alpha value for the RRC Filter. This parameter is available when 3GPP W-CDMA, NADC, or TETRA has been selected as the Radio Std. from the Mode Setup menu.
State Saved: Saved in instrument state.
otherwise Off.
History: Added with firmware revision A.03.00.
[:SENSe]:ACPower:FILTer[:RRC]:ALPHA
[:SENSe]:ACPower:FILTer[:RRC]:ALPHA?
5.6.13 Noise Correction
Pressing Noise Correction turns noise correction on or off. When you set Noise Corr to On, a calibration of the noise floor is performed and used to correct for analyzer noise floor contribution to measurement levels, increasing dynamic range.
Noise Correction is unavailable if signal tracking is on, or when the radio standard is set to NADC or PDC when the measurement method is set to RBW.
State Saved: Saved in instrument state.

Meas Setup (Adjacent Channel Power—ACP)
History: Added with firmware revision A.02.00.
[:SENSe]:ACPower:CORRection:NOISe[:AUTO] OFF|ON|0|1
[:SENSe]:ACPower:CORRection:NOISe[:AUTO]?
Remote Command Notes: The noise correction feature is not available when the radio standard is set to NADC or PDC

Meas Setup (Adjacent Channel Power—ACP)

If ACP has been selected in the Measure menu of the Spectrum Analysis mode, this key displays the appropriate Trace/View menu for the adjacent channel power (ACP) measurement.
Displays menu keys that enable you to set how trace information is stored and displayed.
There is no equivalent remote command for this function.
Pressing Spectrum selects the spectral display of the measurement.
There is no remote command for this function.
Pressing Bars turns the graphic bar display on or off.
There is no remote command for this function.
Trace/View (ACP Measurement)

Pressing Combined selects the measurement to be displayed as a bar graph and spectrum as shown below.

There is no remote command for this function.
Pressing Combined View Units selects the units (dBc or dBm) for the floating numeric displays when Combined is selected.
There is no remote command for this function.
One-button measurements do not change this function. See “Trace” on page 284 for more
Trace/View (ACP Measurement)

information.
5.7.5.1 Trace (1 2 3)
Determines which trace the menu keys will affect. Press Trace 1 2 3 until the number of the desired trace is underlined.
:TRACe[1]|2|3:MODE WRITe|MAXHold|MINHold|VIEW|BLANk
WRITe = Clear Write
MAXHold = Max Hold
MINHold = Min Hold
:TRACe[1]|2|3:MODE?
TRAC:MODE MINH
TRAC:MODE VIEW
TRAC:MODE BLANk
5.7.5.2 Clear Write
See “Clear Write” on page 285 for more information.
5.7.5.3 Max Hold
See “Max Hold” on page 285 for more information.
5.7.5.4 Min Hold
See “Min Hold” on page 285 for more information.
5.7.5.5 View
See “View” on page 286 for more information.
Trace/View (ACP Measurement) · VIEW = View · BLANk = Blank

5.7.5.6 Blank
See “Blank” on page 286 for more information.
Trace/View (ACP Measurement)

Meas Setup (Burst Power)
When Burst Power has been selected in the Measure menu of the Spectrum Analysis Mode, this key displays the appropriate measurement setup menu.
The burst power measurement is an accurate method of determining the average power for the specified burst.
There is no equivalent remote command.
Press Avg Number (On) to specify the number of measurement averages used when calculating the measurement result. The average is displayed at the end of each sweep.
State Saved: Saved in instrument state.
[:SENSe]:BPOWer:AVERage:COUNt
[:SENSe]:BPOWer:AVERage:COUNt?
[:SENSe]:BPOWer:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:BPOWer:AVERage[:STATe]?

Meas Setup (Burst Power)
Press Avg Mode to select the type of termination control used for the averaging function to either Exp or Repeat. This determines the averaging action after the specified number of measurements (average count) is reached.
State Saved: Saved in instrument state.
History: Added with firmware revision A.02.00.
[:SENSe]:BPOWer:AVERage:TCONrol EXPonential|REPeat
[:SENSe]:BPOWer:AVERage:TCONrol?
Example: BPOW:AVG:TCON EXP BPOW:AVG:TCON?
Allows you to specify the type of result averaging to be performed.
[:SENSe]:BPOWer:AVERage:TYPe LPOWer|POWer

Meas Setup (Burst Power)
[:SENSe]:BPOWer:AVERage:TYPe?
Example: BPOW:AVG:TYP LPOWer to select Log type. BPOW:AVG:TYP?
Enables you to set the level above which the mean carrier power calculation is based. The threshold level can be described in dB (Rel) or dBm (Abs).
–60 dB to 0 dB (in relative mode)
[:SENSe]:BPOWer:THReshold
[:SENSe]:BPOWer:THReshold?
[:SENSe]:BPOWer:THReshold:TYPE ABSolute|RELative
[:SENSe]:BPOWer:THReshold:TYPE?
Allows you to select the measurement method.
Default Terminator dB

Meas Setup (Burst Power)
[:SENSe]:BPOWer:METHod THReshold|BWIDth
[:SENSe]:BPOWer:METHod?
Example: BPOW:METH BWID to select burst width as the measurement method.
Sets the burst width parameter to automatic mode (Auto) or manual mode (Man).

Meas Setup (Burst Power)
[:SENSe]:BPOWer:BURSt:WIDTh
[:SENSe]:BPOWer:BURSt:WIDTh?
[:SENSe]:BPOWer:BURSt:AUTO OFF|ON|0|1
[:SENSe]:BPOWer:BURSt:AUTO?
burst width. That functionality is available from the front panel, but the command must be sent with a time value.
Sets the input attenuator to optimize the robustness of the measurement, which is its freedom from errors due to input compression.
State Saved: Saved in instrument state.
History: Added with firmware revision A.02.00.
[:SENSe]:POWer[:RF]:RANGe:AUTO ONCE

Meas Setup (Burst Power)

If Burst Power has been selected in the Measure menu of the Spectrum Analysis mode, this key displays the appropriate Trace/View menu for the burst power measurement.
Displays menu keys that enable you to set how trace information is stored and displayed.
There is no remote command for this key.
Press RF Envelope to view the measurement results as shown in the following figure. The results are updated after each sweep.

Trace/View (Burst Power)

Pressing Combined displays measurement results the same as RF Envelope, but has a blue bar between the markers to indicate the measured output power level. The actual measure output power is displayed at the bottom of the bar. Refer to the following figure.

One-button measurements do not change this function. See “Trace” on page 284 for more information.
5.9.3.1 Trace (1 2 3)
Determines which trace the menu keys will affect. Press Trace 1 2 3 until the number of the desired trace is underlined.
:TRACe[1]|2|3:MODE WRITe|MAXHold|MINHold|VIEW|BLANk
WRITe = Clear Write
MAXHold = Max Hold
MINHold = Min Hold
Trace/View (Burst Power) · VIEW = View

:TRACe[1]|2|3:MODE?
TRAC:MODE MINH
TRAC:MODE VIEW
TRAC:MODE BLANk
5.9.3.2 Clear Write
See “Clear Write” on page 285 for more information.
5.9.3.3 Max Hold
See “Max Hold” on page 285 for more information.
5.9.3.4 Min Hold
See “Min Hold” on page 285 for more information.
5.9.3.5 View
See “View” on page 286 for more information.
5.9.3.6 Blank
See “Blank” on page 286 for more information.
Trace/View (Burst Power) · BLANk = Blank

Trace/View (Burst Power)

Meas Setup (Complementary Cumulative Distribution Function—CCDF)
When Power Stat CCDF has been selected in the Measure menu of the Spectrum Analysis mode, this key displays the appropriate measurement setup menu for the complementary cumulative distribution function (CCDF).
Power Complementary cumulative distribution function (CCDF) curves characterize the higher-level power of the signal. It provides the distribution of peak-to-average power ratios versus probability. A CCDF curve is defined by how much time the waveform spends at or above the specified power level. The percent of time the signal spends at or above this level defines the probability for that particular power level.
There is no equivalent remote command.
Press Meas BW to set the measurement 3 dB bandwidth. The range is 10 kHz to 8 MHz, using the available RBW filters.
State Saved: Saved in instrument state.
Factory Preset: Defined by the radio standard selected.
Knob Increment: Steps through available Res BW filters.
Step Key Increment: Steps through available Res BW filters.
[:SENSe]:PSTatistic:BANDwidth|BWIDth
[:SENSe]:PSTatistic:BANDwidth|BWIDth?
Remote Command Notes: The entered value is rounded to the nearest Res BW filter frequency.

Meas Setup (Complementary Cumulative Distribution Function—CCDF)
PST:BAND? or PST:BWID?
Press Counts to set the accumulated number of sampling points for data acquisition. The range is 1 kpoints (kilopoint–1000 points) to 2.000000 Gpoints (Giga point–109 point) with 1 kpoint resolution. While this key is activated, enter a value from the numeric keypad by terminating with one of the unit keys shown.
State Saved: Saved in instrument state.
[:SENSe]:PSTatistic:COUNts
[:SENSe]:PSTatistic:COUNts?
PST:COUNt?
Press Meas Interval to specify the time interval over which the measurement is made. The minimum setting is 100 µs. The maximum setting depends on the measurement bandwidth setting but never exceeds 600 ms. The resolution is 1 µs.
State Saved: Saved in instrument state.

Meas Setup (Complementary Cumulative Distribution Function—CCDF)
[:SENSe]:PSTatistic:SWEep:TIME
[:SENSe]:PSTatistic:SWEep:TIME?
5.10.4 Optimize Ref Level
Sets the input attenuator to optimize the robustness of the measurement, which is its freedom from errors due to input compression.
State Saved: Saved in instrument state.
[:SENSe]:POWer[:RF]:RANGe:AUTO ONCE

Meas Setup (Complementary Cumulative Distribution Function—CCDF)

Display (Complementary Cumulative Distribution Function—CCDF)
Press Display to access menu keys that allow you to configure parameters for the CCDF measurement.
Expands the measurement window horizontally over the entire instrument display. Pressing any other key that results in a new menu cancels the full screen function.
Any error occurring turns full screen mode off.
State Saved: Not saved in state.
Press Store Ref Trace to copy the currently measured curve as the user-definable reference trace. The captured data will remain until the other mode is chosen. Pressing this key refreshes the reference trace.
:CALCulate:PSTatistic:STORe:REFerence

Display (Complementary Cumulative Distribution Function—CCDF)
Press Ref Trace to toggle the reference trace display function between On and Off.
State Saved: Saved in instrument state.
:DISPlay:PSTatistic:REFerence[:STATe] OFF|ON|0|1
:DISPlay:PSTatistic:REFerence[:STATe]?
5.11.4 Gaussian Trace
Press Gaussian Line to toggle the Gaussian line display function between On and Off.
State Saved: Saved in instrument state.
:DISPlay:PSTatistic:GAUSsian[:STATe] OFF|ON|0|1
:DISPlay:PSTatistic:GAUSsian[:STATe]?
5.11.5 Preferences
Displays the menu key that enables you to turn the graticule and annotation on or off.
State Saved: Saved in instrument state.

Display (Complementary Cumulative Distribution Function—CCDF)

Display (Complementary Cumulative Distribution Function—CCDF)
5.11.5.1 Graticule
Pressing Graticule turns the display graticules On or Off.

SPAN X Scale (Complementary Cumulative Distribution Function—CCDF)
Activates the Span function for the CCDF measurement and displays the menu of span functions.
See the Scale/Div command below.
Allows you to enter a numeric value to change the horizontal display sensitivity by setting the decibel units per horizontal graticule division.
State Saved: Saved in instrument state.
:DISPlay:PSTatistic:XSCale
:DISPlay:PSTatistic:XSCale?
Example DISP:PST:XSC 10 dB

SPAN X Scale (Complementary Cumulative Distribution Function—CCDF)

Marker (Complementary Cumulative Distribution Function—CCDF)
Displays the Marker control menu for the CCDF measurement. All markers are off until either Normal or Delta is pressed. There are three control modes for the markers:
State Saved: The control mode for each marker, as well as the position of each marker, is saved in instrument state.
Factory Preset: All Off.
There is no equivalent remote command. These commands control the three marker modes. Refer to the Normal, Delta, and Off key descriptions for additional information.
:CALCulate:PSTatistic:MARKer[1]|2|3|4:MODE NORMal|DELTa|OFF
:CALCulate:PSTatistic:MARKer[1]|2|3|4:MODE?
“Marker X Position” on page 395
“Marker Y Position (Remote Command Only)” on page 396
“Marker Maximum and Minimum (Remote Command Only)” on page 397
Example: CALC:PST:MARK:MODE DELT activates a delta marker (marker 1) at the 0 dB point of your measurement (yellow line).

Marker (Complementary Cumulative Distribution Function—CCDF)
Selects one of the four possible markers. Once a marker is selected, it can be set to any of the control modes Normal, Delta, or Off.
State Saved: The number of the selected marker is saved in instrument state.
:CALCulate:PSTatistic:MARKer[1]|2|3|4[:STATe] OFF|ON|0|1
:CALCulate:PSTatistic:MARKer[1]|2|3|4[:STATe]?
Sets or queries the state of a marker. Setting a marker to state ON or 1 selects that marker. Setting a marker which is OFF to state ON or 1 puts it in Normal mode and places it at the center of the display. Setting a marker to state OFF or 0 selects that marker and turns it off. The response to the query will be 0 if OFF, 1 if ON.
Example: CALC:PST:MARK2 ON selects marker 2.
CALC:PST:MARK:STAT ON will not modify a marker that is already on.
Sets the control mode for the selected marker to Normal. If the marker is off, a single marker is activated at the 0 dB point of the selected trace (see “Marker Trace” on page 394). You can then adjust the selected marker using the numeric keypad, knob, or keys.
See “Marker” on page 177 for the mode command.
Example: CALC:PST:MARK:MODE NORM selects marker 1 and sets it to Normal.
Sets the control mode for the selected marker to Delta. In Delta mode the display shows the difference between the active (Delta) marker and a reference marker. When Delta mode is selected the reference marker is placed at the current marker position, unless the marker was OFF, in which case both the active marker and the reference marker are placed at 0 dB point of the selected trace (see “Marker Trace” on page 394). You can adjust the trace point of the active delta marker. Annotation in the active function block and in the upper-right corner of the display indicates the power ratio (dB) difference and probability (percentage) difference of the two markers.

Marker (Complementary Cumulative Distribution Function—CCDF)

Marker (Complementary Cumulative Distribution Function—CCDF)
See “Marker” on page 177 for the mode command.
Example: CALC:PST:MARK3:MODE DELT selects marker 3 as a delta marker and places a reference marker at the 0 dB point on the specified trace (see “Marker Trace” on page 394). If marker 3 is OFF it places both the active and the reference markers at the 0 dB point of the specified trace.
Turns off the selected marker. In addition, Off also turns off functions related to the selected marker.
See “Select Marker” on page 180 for the command to select the control mode.
Example:
CALC:PST:MARK3:STAT OFF selects marker 3 and sets it to Off.
Chooses which trace the selected marker will be placed on. You can pick Measured, Gaussian, or Reference.
State Saved: The Marker Trace for each marker is saved in instrument state.
:CALCulate:PSTatistic:MARKer[1]|2|3|4:TRACe MEASured|GAUSsian|REFerence
Puts the marker on the specified trace.
:CALCulate:PSTatistic:MARKer[1]|2|3|4:TRACe?
The query returns the type of the trace on which the marker currently resides.
Example: CALC:PST:MARK1:TRAC GAUSsian places marker 1 on the Gaussian trace.
Turns off all markers.

Marker (Complementary Cumulative Distribution Function—CCDF)
:CALCulate:PSTatistic:MARKer[1]|2|3|4:AOFF
Example: CALC:PST:MARK:AOFF turns off all markers.
Sets the marker X position (horizontal) to a specified point on the X axis in the current X-axis units (dB). If the value chosen would place the marker off screen, the marker will be placed at the left or right side of the display, on the specified trace (see “Marker Trace” on page 394). If the marker is off, this command has no effect.
Key Notes: When a marker is active, you may use the knob, numeric keypad, or keys to specify the horizontal location of the marker.
:CALCulate:PSTatistic:MARKer[1]|2|3|4:X
Sets the marker X position in the current X-axis units (dB).
:CALCulate:PSTatistic:MARKer[1]|2|3|4:X?
Queries the marker X position in the current x-axis units. For the response to be valid, the marker must be on.
:CALCulate:PSTatistic:MARKer[1]|2|3|4:X:POSition
:CALCulate:PSTatistic:MARKer[1]|2|3|4:X:POSition?
This defines marker X position in display points from 0 to 50 dB.
Example: CALC:PST:MARK:X 30 dB activates a normal marker (marker 1) at the 30 dB location on the specified trace (see “Marker Trace” on page 394).
CALC:PST:MARK2:X:POS 10 selects marker 2 and moves it to 1 dB location on the specified trace.

Marker (Complementary Cumulative Distribution Function—CCDF)
Displays the marker Y position (vertical) of the specified trace (see “Marker Trace” on page 394) in the current Y-axis units (percent).

Marker (Complementary Cumulative Distribution Function—CCDF)
:CALCulate:PSTatistic:MARKer[1]|2|3|4:Y?
Queries the marker Y value or delta in the current y axis units. The marker must be ON for the response to be valid.
Example: CALC:PST:MARK1:Y?
Moves the marker to the maximum or minimum detected amplitude value on the display. If the marker is off, this command has no effect.
:CALCulate:PSTatistic:MARKer[1]|2|3|4:MAXimum
:CALCulate:PSTatistic:MARKer[1]|2|3|4:MINimum

Marker (Complementary Cumulative Distribution Function—CCDF)

Meas Setup (Channel Power—CHP)
When the channel power measurement has been selected in the Measure key menu of the Spectrum Analysis Mode, this key displays the appropriate measurement setup menu.
The Channel Power measurement measures the power and power spectral density in the channel bandwidth that you specify. One marker pair on the display indicates the edges of the channel bandwidth. The center frequency, reference level, and channel bandwidth must be set by the user.
Dependencies/ Couplings: Menu changes depending on the Mode and Measurement selected.
There is no equivalent remote command.
To specify the number of measurement averages used when calculating the measurement result set Avg Number to On. The average will be displayed at the end of each sweep. Setting Avg Number to Off disables measurement averaging.
State Saved: Saved in instrument state.
[:SENSe]:CHPower:AVERage:COUNt
[:SENSe]:CHPower:AVERage:COUNt?
[:SENSe]:CHPower:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:CHPower:AVERage[:STATe]?

Meas Setup (Channel Power—CHP)
CHP:AVER OFF
CHP:AVER?
5.14.2 Avg Mode
Press Avg Mode to select the type of termination control used for the averaging function to either Exp or Repeat. This determines the averaging action after the specified number of measurements (average count) is reached.
State Saved: Saved in instrument state.
[:SENSe]:CHPower:AVERage:TCONrol EXPonential|REPeat
[:SENSe]:CHPower:AVERage:TCONrol?
5.14.3 Integ BW
Press Integ BW to specify the range of integration used in calculating the power in the channel, for example. set the main (center) channel bandwidth. Note that the integration bandwidth is displayed on the trace as two markers connected by an arrow. Be sure the Span of the instrument is set between 1 and 10 times the integration bandwidth.
Key Notes: If RRC Filter is on, the actual integration bandwidth used is the displayed integration bandwidth multiplied by (1 + “Filter Alpha”).
State Saved: Saved in instrument state.

Meas Setup (Channel Power—CHP)
Factory Preset: 2 MHz, or as defined by the selected radio standard.
Step Key Increment: 1,1.5, 2, 3, 5, 7.5, 10...
Range: 100 Hz to maximum span supported by your analyzer.
[:SENSe]:CHPower:BANDwidth|BWIDth:INTegration
[:SENSe]:CHPower:BANDwidth|BWIDth:INTegration?
5.14.4 Chan Pwr Span
Press Chan Pwr Span to set the analyzer span for the channel power measurement. When the RRC Filter is set to On, the lower limit for the span is .
1 alpha + ( ) Integration BW ×
State Saved: Saved in instrument state.
Factory Preset: 3 MHz, or as defined by the selected radio standard.
Knob Increment: 1, 1.5, 2, 3, 5, 7.5, 10...
Range: Current integration bandwidth to 10 times the integration bandwidth or span of your analyzer.
[:SENSe]:CHPower:FREQuency:SPAN
[:SENSe]:CHPower:FREQuency:SPAN?

Meas Setup (Channel Power—CHP)
Sets the input attenuator to optimize the robustness of the measurement, which is its freedom from errors due to input compression.
State Saved: Not saved.
[:SENSe]:POWer[:RF]:RANGe:AUTO ONCE
Pressing RRC Filter turns the Root Raised Cosine filter on or off. This filter is the type specified in the TETRA and 3GPP W-CDMA standards. This parameter is only available when either 3GPP W-CDMA or TETRA has been selected as the Radio Std from the Mode Setup menu. If 3GPP W-CDMA is selected, the roll-off value (alpha) for the filter will be initially set to 0.22. T is set to 1/RefBW (default 260 ns). If TETRA is selected, the roll-off will be initially set to 0.35. T is set to 1/RefBW (default 55.56 µs). The roll-off value can be changed using Filter Alpha.
State Saved: Saved in instrument state.
History: Added with firmware revision A.03.00.
[:SENSe]:CHPower:FILTer[:RRC][:STATe] OFF|ON|0|1
[:SENSe]:CHPower:FILTer[:RRC][:STATe]?
5.14.7 Filter Alpha
Press RRC Filter to input the alpha value for the RRC Filter. This parameter is only available when either 3GPP W-CDMA or TETRA has been selected as the Radio Std. from the Mode Setup menu.

Meas Setup (Channel Power—CHP)
State Saved: Saved in instrument state.
Factory Preset: 0.22 when W-CDMA is selected, 0.35 for TETRA, otherwise Off.
History: Added with firmware revision A.03.00.

Meas Setup (Channel Power—CHP)
[:SENSe]:CHPower:FILTer[:RRC]:ALPHA
[:SENSe]:CHPower:FILTer[:RRC]:ALPHA?

Trace/View (Channel Power Measurement)
If Channel Power has been selected in the Measure menu of the Spectrum Analysis mode, this key displays the appropriate Trace/View menu for the channel power measurement.
Displays menu keys that enable you to set how trace information is stored and displayed.
There is no equivalent remote command for this function.
Pressing Spectrum selects the spectral display of the measurement.
There is no remote command for this function.
Pressing Combined selects the measurement to be displayed as a bar graph and spectrum.
There is no remote command for this function.
One-button measurements do not change this function. See “Trace” on page 284 for more information.

Trace/View (Channel Power Measurement)
5.15.3.1 Trace (1 2 3)
Determines which trace the menu keys will affect. Press Trace 1 2 3 until the number of the desired trace is underlined.
:TRACe[1]|2|3:MODE WRITe|MAXHold|MINHold|VIEW|BLANk
WRITe = Clear Write
MAXHold = Max Hold
MINHold = Min Hold
:TRACe[1]|2|3:MODE?
TRAC:MODE MINH
TRAC:MODE VIEW
TRAC:MODE BLANk
5.15.3.2 Clear Write
See “Clear Write” on page 285 for more information.
5.15.3.3 Max Hold
See “Max Hold” on page 285 for more information.
5.15.3.4 Min Hold
See “Min Hold” on page 285 for more information.
5.15.3.5 View
See “View” on page 286 for more information.
VIEW = View · BLANk = Blank

Trace/View (Channel Power Measurement)
5.15.3.6 Blank
See “Blank” on page 286 for more information.

Trace/View (Channel Power Measurement)

Meas Setup (Harmonic Distortion)
When the harmonic distortion measurement has been selected in the Measure key menu of the Spectrum Analysis Mode, this key displays the appropriate measurement setup menu.
Saved State: Not saved.
Press Avg Number (On) to specify the number of measurement averages used when calculating the measurement result. The average will be displayed at the end of each sweep.
State Saved: Saved in instrument state.
[:SENSe]:HARMonics:AVERage:COUNt
[:SENSe]:HARMonics:AVERage:COUNt?
[:SENSe]:HARMonics:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:HARMonics:AVERage[:STATe]?
5.16.2 Avg Mode
Press Avg Mode to select the type of termination control used for the averaging function to either Exp or Repeat. This determines the averaging action after the specified number of measurements (average count) is reached.

Meas Setup (Harmonic Distortion)
State Saved: Saved in instrument state.
History: Added with firmware revision A.02.00.
[:SENSe]:HARMonics:AVERage:TCONrol EXPonential|REPeat
[:SENSe]:HARMonics:AVERage:TCONrol?
5.16.3 Harmonics
Harmonics indicates the number of harmonics to measure before computing the total harmonic distortion. The minimum number is 2 (only the fundamental and second harmonic will be measured). The maximum number is 10.
State Saved: Saved in instrument state.
[:SENSe]:HARMonics:NUMBer
[:SENSe]:HARMonics:NUMBer?
Sets the sweep time used to measure each harmonic. The value is set to 200 divided by the resolution bandwidth, or 10 ms, whichever is greater when the measurement is started. This sweep time is used only for measuring harmonics. The analyzer sweep time before the measurement was started is used for finding the fundamental.

Meas Setup (Harmonic Distortion)
State Saved: Saved in instrument state.

Meas Setup (Harmonic Distortion)
[:SENSe]:HARMonics:SWEeptime:STATe OFF|ON|0|1
[:SENSe]:HARMonics:SWEeptime:STATe?
[:SENSe]:HARMonics:SWEeptime
[:SENSe]:HARMonics:SWEeptime?
Remote Command Notes: ON is equal to automatic (Auto). OFF is equal to manual (Man).
HARM:SWE 100 ms
Press Range Table to specify whether the range table is to be used or not. When Range Table is set to Off the harmonics are measured. When Range Table is set to On the first active range is considered to be the fundamental and the subsequent active ranges are used as the harmonics. At least two ranges must be active if Range Table is set to On.
Factory Preset: See Table 5-1 on page 414.
[:SENSe]:HARMonics:RTABle:STATe OFF|ON|0|1
[:SENSe]:HARMonics:RTABle:STATe?

Meas Setup (Harmonic Distortion)
Enables you to enter the settings for up to 10 ranges, either using the instrument front panel keys or remotely. Upon entering the range table, the measurement stops, then the analyzer is set to a constantly sweeping state. The analyzer is then set to the current values for range 1 (whether range 1 is on or off). If a range is currently off, the values in the range table for that range are replaced with --- (see Range 5 in the illustration) to indicate this range is currently inactive.

Saved State: All values for all ranges are saved in instrument state.

Meas Setup (Harmonic Distortion)
Span, Res BW Mode(Man), Video BW(Man) is set to 2 kHz, and Sweep Time Mode(Auto) is set to 10 ms. The table below defines the other default parameter settings for the measurement.
Table 5-1 Range Table Default Settings
Range Center Freq (GHz)
1 1.000 100.0
2 2.000 200.0
3 3.000 300.0
4 4.000 430.0
5 5.000 510.0
6 6.000 620.0
7 7.000 750.0
8 8.000 820.0
9 9.000 910.0
10 10.00 1000.0
There is no equivalent remote command.
5.16.6.1 Range
Selects a range and updates the values on the other Range Table keys so that they reflect the settings for the selected range. If Range is set to On it is used as part of the measurement; when set to Off it is excluded. A range is made up of the next parameters.
Factory Preset: See Table 5-1 on page 414.
Res BW (kHz)

Meas Setup (Harmonic Distortion)
This parameter can send up to 10 values. The location in the list sent corresponds to the range the value is associated with. Missing values are not permitted. For example, if you want to change values 2 and 6, you must send all values up to 6. Subsequent values remain as they were.
[:SENSe]:HARMonics:RANGe[:LIST]:STATe OFF|ON|0|1
The query for this parameter will always return 20 values.
[:SENSe]:HARMonics:RANGe[:LIST]:STATe?
5.16.6.2 Center Freq
Used to set the center frequency of the analyzer.
Factory Preset: See Table 5-1 on page 414.
Step Key Increment: 1, 1.5, 2, 3, 5, 7.5, 10, ...
Range: Frequency range of your analyzer.
This parameter can send up to 10 values. The location in the list sent corresponds to the range the value is associated. Missing values are not permitted, for example, if you want to change values 2 and 6, you must send all values up to 6. Subsequent values will remain as they were.
[:SENSe]:HARMonics:RANGe[:LIST]:FREQuency
The query for this parameter will always return 10 values.
[:SENSe]:HARMonics:RANGe[:LIST]:FREQuency?
5.16.6.3 Span
Span is used to set the span of the analyzer.
Factory Preset: See Table 5-1 on page 414.

Meas Setup (Harmonic Distortion)
Knob Increment: Steps through the available spans.
Step Key Increment: Steps through the available spans.
Range: Span of your analyzer.
This parameter can send up to 10 values. The location in the list sent corresponds to the range the value is associated with. Missing values are not permitted, for example, if you want to change values 2 and 6, you must send all values up to 6. Subsequent values will remain as they were.
[:SENSe]:HARMonics:RANGe[:LIST]:SPAN
[:SENSe]:HARMonics:RANGe[:LIST]:SPAN?
5.16.6.4 Res BW
Res BW is used to set the resolution bandwidth mode of the analyzer. If the range is zero span and Res BW is set to Auto, the analyzer determines the optimum setting. Res BW set to Man enables you to determine the setting.
Factory Preset: See Table 5-1 on page 414.
Knob Increment: Steps through the available resolution bandwidth filters.
Step Key Increment: Steps through the available resolution bandwidth filters.
Range: Resolution bandwidth range of your analyzer.
This parameter can send up to 10 values. The location in the list sent corresponds to the range the value is associated with. Missing values are not permitted, for example, if you want to change values 2 and 6, you must send all values up to 6. Subsequent values will remain as they were.
[:SENSe]:HARMonics:RANGe[:LIST]:BWIDth|BANDwidth[:RESolution]:AUTO OFF|ON|0|1 selects the mode.
[:SENSe]:HARMonics:RANGe[:LIST]:BWIDth|BANDwidth[:RESolution]
The query for this parameter always returns 10 values.
[:SENSe]:HARMonics:RANGe[:LIST]:BWIDth|BANDwidth[:RESolution]:AUTO?

Meas Setup (Harmonic Distortion)
[:SENSe]:HARMonics:RANGe[:LIST]:BWIDth|BANDwidth[:RESolution]?

Meas Setup (Harmonic Distortion)
5.16.6.5 Video BW
Video BW is used to set the video bandwidth of the analyzer. When Auto is selected the analyzer determines the optimum setting, while Man enables you to determine the setting.
Knob Increment: Steps through the available video bandwidth filters.
Step Key Increment: Steps through the available video bandwidth filters.
Range: Video bandwidth range of your analyzer.
This parameter can send up to 10 values. The location in the list sent corresponds to the range the value is associated with. Missing values are not permitted, for example, if you want to change values 2 and 6, you must send all values up to 6. Subsequent values will remain as they were.
[:SENSe]:HARMonics:RANGe[:LIST]:BWIDth|BANDwidth:VIDeo:AUTO OFF|ON|0|1 selects the mode.
[:SENSe]:HARMonics:RANGe[:LIST]:BWIDth|BANDwidth:VIDeo
The query for this parameter will always return 10 values.
[:SENSe]:HARMonics:RANGe[:LIST]:BWIDth|BANDwidth:VIDeo:AUTO?
[:SENSe]:HARMonics:RANGe[:LIST]:BWIDth|BANDwidth:VIDeo?
5.16.6.6 Sweep Time
Sweep Time is used to set the sweep time of the analyzer. If the range is zero span and Sweep Time is set to Auto, the sweep time will be 200 divided by the resolution bandwidth or 10 ms, whichever is greater. If the range is non zero span and Sweep Time is set to Auto, the analyzer determines the optimum setting. Sweep Time set to Man enables you to determine the setting.
Step Key Increment: 1, 1.5, 2, 3, 5, 7.5, 10, ...
Range: Sweep time range of your analyzer.

Meas Setup (Harmonic Distortion)
This parameter can send up to 10 values. The location in the list sent corresponds to the range the value is associated with. Missing values are not permitted, for example, if you want to change values 2 and 6, you must send all values up to 6. Subsequent values will remain as they were.
[:SENSe]:HARMonics:RANGe[:LIST]:SWEep:TIME:AUTO OFF|ON|0|1 selects the mode.
[:SENSe]:HARMonics:RANGe[:LIST]:SWEep:TIME:
The query for this parameter will always return 10 values.
[:SENSe]:HARMonics:RANGe[:LIST]:SWEep:TIME:AUTO?
[:SENSe]:HARMonics:RANGe[:LIST]:SWEep:TIME?
5.16.6.7 Auto Fill Range Table
Auto Fill Range Table is used to automatically setup the range table based on the current range center frequency and span. When selected, the current center frequency is taken as the fundamental and the frequency in range one is set to this value. Each of the subsequent range center frequency parameters are set to the appropriate integer multiple of the fundamental frequency. The span for range one is set to that of the current range. Each of the subsequent range span parameters are set to the appropriate integer multiple of the fundamental span. This means that if the fundamental range is zero span, all the ranges will be zero span. If the range span is zero, Res BW, Video BW, and Sweep Time are set to the current base instrument settings. If the range span is non-zero, Res BW, Video BW, and Sweep Time for all harmonics are set to the same settings as for the fundamental. Span is set to zero for all harmonics and the fundamental.
Step Key Increment: 1, 1.5, 2, 3, 5, 7.5, 10, ...
Range: Sweep time range of your analyzer.
[:SENSe]:HARMonics:RTABle:FILL
Sets the input attenuator to optimize the robustness of the measurement, which is its freedom from errors due to input compression.

Meas Setup (Harmonic Distortion)
potential errors due to VSWR.
State Saved: Saved in instrument state.
History: Added with firmware revision A.02.00.
[:SENSe]:POWer[:RF]:RANGe:AUTO ONCE

If Harmonic Distortion has been selected in the Measure menu of the Spectrum Analysis mode, this key displays the appropriate Trace/View menu to view the harmonic measurement results.
Displays menu keys to view the harmonic measurement results in two views, Harmonic and Harmonic & THD.
Added with firmware revision A.02.00
Press Harmonics to view the measurement results in numeric form.
Press Harmonics & THD to view the measurement results in numeric form, as well as the total harmonic distortion (THD) displayed as a percentage and as a dB value.

Trace/View (Harmonics)

Trace/View (Harmonics)

Meas Setup (Intermod (TOI))
When Intermod (TOI) measurement has been selected in the Measure menu of the Spectrum Analysis Mode, this key displays the appropriate measurement setup menu for third order intercept (TOI).
The TOI measurement computes and displays the output intercept point (IP3), and places markers upon the trace to indicate the measured signals and third-order products.
Dependencies/ Couplings: Menu changes depending on the Mode and Measurement selected.
There is no equivalent remote command.
Set Avg Number to On to specify the number of measurement averages used when calculating the measurement result. The average will be displayed at the end of each sweep. Setting Avg Number to Off disables measurement averaging.
State Saved: Saved in instrument state.
[:SENSe]:TOIN:AVERage:COUNt
[:SENSe]:TOIN:AVERage:COUNt?
[:SENSe]:TOIN:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:TOIN:AVERage[:STATe]?

Meas Setup (Intermod (TOI))

Meas Setup (Intermod (TOI))
Press Avg Mode to select the type of termination control used for the averaging function to either Exp or Repeat. This determines the averaging action after the specified number of measurements (average count) is reached.
State Saved: Saved in instrument state.
[:SENSe]:TOIN:AVERage:TCONrol EXPonential|REPeat
[:SENSe]:TOIN:AVERage:TCONrol?
5.18.3 TOI Span
Press Span to specify the frequency span in which intermodulation products are measured.
Key Note: If you modify the value of Span in the base instrument Span menu, the value in the Meas Setup menu will be updated to reflect the new value and the measurement will restart if it is running.
Factory Preset: 15 MHz, or as defined by the radio standard selected.
[:SENSe]:TOIN:FREQuency:SPAN
[:SENSe]:TOIN:FREQuency:SPAN?

Meas Setup (Intermod (TOI))
5.18.4 Max Mixer Lvl
Enables you to set the relationship between the highest signal that can be displayed (the reference level) and the input attenuation. The relationship applies whenever the Attenuation is set to Auto. The relationship is that the attenuation is given by reference level minus the max mixer level. For example, as the reference level changes, the input attuanator changes to ensure that a signal at the reference level does not exceed the Max Mixer Lvl setting.
When the TOI measurement is on, this key controls the maximum mixer level, independent of the previous setting of Max Mixer Lvl located under Amplitude. When the TOI measurement is Off, the previous maximum Max Mixer Lvl is restored.
Setting Max Mixer Lvl to Auto sets the maximum mixer level to −30 dBm.
Range:
[:SENSe]:TOIN:FREQuency:MIXer:RANGe[:UPPer]
[:SENSe]:TOIN:FREQuency:MIXer:RANGe[:UPPer]?
[:SENSe]:TOIN:FREQuency:MIXer:RANGe:AUTO OFF|ON|0|1
[:SENSe]:TOIN:FREQuency:MIXer:RANGe:AUTO?
TOIN:FREQ:MIX:RANG:AUTO 1
TOIN:FREQ:MIX:RANG:AUTO?
5.18.5 Optimize Ref Level
Sets the reference level to place the tones near the top of the display. When input attenuation is set to Auto, pressing Optimize Ref Level sets the input attenuation according to the auto rules described in Max Mixer Lvl, based on this reference level. You may want to trade off analyzer noise and analyzer-induced distortion by manually changing the attenuation setting from this starting point.

Meas Setup (Intermod (TOI))
State Saved: Saved in instrument state.

Meas Setup (Intermod (TOI))
[:SENSe]:POWer[:RF]:RANGe:AUTO ONCE

Meas Setup (Multi-Carrier Power—MCP)
If the MCP measurement has been selected in the Measure menu of the Spectrum Analysis mode, this key displays the MCP measurement setup menu.
The Multi-Carrier Power measurement is a measure of the power in two or more transmit channels and of the power that leaks into their adjacent transmit channels. The results reported are similar to the adjacent channel power measurement, but the setup is different to allow for two or more carriers present.
Dependencies/ Couplings: • Parameter defaults change depending on the Radio Standard selected.
There is no equivalent remote command.
Pressing Avg Number to On enables you to specify the number of measurements that will be averaged when calculating the measurement result. The average will be displayed at the end of each sweep. Setting Avg Number to Off disables the measurement averaging.
State Saved: Saved in instrument state.
Remote Commands:
[:SENSe]:MCPower:AVERage:COUNt
[:SENSe]:MCPower:AVERage:COUNt?
[:SENSe]:MCPower:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:MCPower:AVERage[:STATe]?

Meas Setup (Multi-Carrier Power—MCP)
MCP:AVER?
5.19.2 Avg Mode
Press Avg Mode to select the type of termination control used for the averaging function as either Exp or Repeat. This determines the averaging action after the specified number of measurements (average count) is reached.
State Saved: Saved in instrument state.
[:SENSe]:MCPower:AVERage:TCONrol EXPonential|REPeat
[:SENSe]:MCPower:AVERage:TCONrol?
5.19.3 Carrier Setup
Accesses the Carrier Setup and Configure Carriers menus that allow you to define the various parameters for each carrier.
History: Added with firmware revision A.03.00.
There is no equivalent remote command.

Meas Setup (Multi-Carrier Power—MCP)
5.19.3.1 Carriers
Press Carriers to specify the number of carriers to be measured.
History: Added with firmware revision A.03.00.
[:SENSe]:MCPower:CARRier:COUNt
[:SENSe]:MCPower:CARRier:COUNt?
5.19.3.2 Ref Carrier
Press Ref Carrier (Man) to specify the carrier (identified by a numeric position) from which all relative power measurements will be made. When Ref Carrier is set to Auto, the analyzer selects the carrier with the highest power as the reference.
Key Notes: The carrier must have power present in order to be assigned as a reference carrier frequency. If you change the reference carrier’s power present value from “yes” to “no”, the next carrier to the left (or to the right, if there are none to the left) will be assigned as the reference carrier. If it currently has power present set to No, this will change to Yes. This is also true when there are only two carriers with a “yes” value. Refer to the “Carrier Pwr Present” key description for more information on setting the carrier power present value.
History: Added with firmware revision A.03.00.
[:SENSe]:MCPower:RCARrier:AUTO OFF|ON|0|1
[:SENSe]:MCPower:RCARrier:AUTO?
[:SENSe]:MCPower:RCARrier

Meas Setup (Multi-Carrier Power—MCP)
[:SENSe]:MCPower:RCARrier?
Remote Command Notes: Refer to “Key Notes:” above.
MCP:RCAR?

Meas Setup (Multi-Carrier Power—MCP)
5.19.3.3 Ref Carrier Freq
Press Ref Carrier Freq (Man) to select the reference carrier frequency for this measurement. The center frequency is then calculated using the algorithm below:
Pressing Ref Carrier Freq (Auto) distributes the carriers evenly around the current center frequency. The reference carrier frequency is then calculated using the algorithm below:
The above procedure ensures carrier visibility on the analyzer display.
Dependencies/ Couplings: The reference carrier can be re-assigned by implementing changes to carrier power present values.
State Saved: Saved in instrument state.
“algorithm” above when defining the Ref Carrier Freq(Auto) key.
Step Key Increment: If CF Step(Auto) is selected: span/10. If CF Step(Man) is selected: CF Step
History Added with firmware revision A.03.00.
[:SENSe]:MCPower:RCFRequency
[:SENSe]:MCPower:RCFRequency?
[:SENSe]:MCPower:RCFRequency:AUTO OFF|ON|0|1
[:SENSe]:MCPower:RCFRequency:AUTO?
Default Terminator GHz

Meas Setup (Multi-Carrier Power—MCP)
5.19.3.4 Configure Carriers
Accesses the Config Carriers menu that allows further definition of each carrier.
History: Added with firmware revision A.03.00.
There is no equivalent remote command.
5.19.3.4.1 Carrier
Selects the carrier number you wish to configure.
Dependencies/ Couplings: All keys available on the Configure Carrier key menu are coupled to this key.
State Saved: Saved in instrument state.
History: Added with firmware revision A.03.00.
There is no equivalent remote command.
5.19.3.4.2 Carrier Pwr Present
Press Carrier Pwr Present (Yes) to specify carriers which have power present. First, press the Carrier key and select the carrier number you wish to define, using the Step Keys ( ), the knob, or the numeric keypad. (The carrier number selected is shown on the Carrier key.) Then toggle the Carrier Pwr Present key to indicate either yes or no.
If a carrier is defined as having no power present, the power displayed will be relative to the reference carrier. If a carrier is defined as having power present, the absolute power will be displayed.
Carriers with power present are included in the total carrier power result.

Meas Setup (Multi-Carrier Power—MCP)
Dependencies/ Couplings: This key assigns a value (yes or no) to the carrier number displayed on the Carrier key.
State Saved: Saved in instrument state.
History: Added with firmware revision A.03.00.
[:SENSe]:MCPower:CARRier:LIST:PPResent YES|NO
[:SENSe]:MCPower:CARRier:LIST:PPResent?
Remote Command Notes:
5.19.3.4.3 Carrier Width
Press Carrier Width to specify the width of each carrier including carriers with no power present. There is a corresponding one-to-one relationship between each carrier width value and the specified number of carriers. First, press the Carrier key and select the carrier number you wish to define, using the Step Keys ( ), the knob, or the numeric keypad. (The carrier number selected is shown on the Carrier key.) Then press the Carrier Width key and enter the width using the numeric keypad.
Key Notes: The Carrier key determines which carrier width you are defining.
State Saved: Saved in instrument state.

Meas Setup (Multi-Carrier Power—MCP)
Step Key Increment: If CF Step(Auto) is selected: span/10. If CF Step(Man) is selected: CF Step
History: Added with firmware revision A.03.00.

Meas Setup (Multi-Carrier Power—MCP)
[:SENSe]:MCPower:CARRier:LIST:WIDTh
[:SENSe]:MCPower:CARRier:LIST:WIDTh?
Remote Command Notes:
5.19.3.4.4 Carrier IntegBW
Press Carrier IntegBW to define the integration bandwidth used to calculate the power in the carriers.
State Saved: Saved in instrument state.
radio standard selected. If Mode Setup, Radio Std, 3GPP W-CDMA is selected and the RRC filter is on, the actual integration bandwidth used will be the displayed integration bandwidth multiplied by (1 + filter alpha).
Resolution:
Step Key Increment: If CF Step(Auto) is selected: span/10. If CF Step(Man) is selected: CF Step
History: Added with firmware revision A.03.00.
Default Terminator Hz

Meas Setup (Multi-Carrier Power—MCP)
[:SENSe]:MCPower:CARRier:LIST:BANDwidth|BWIDth:[INTegration]
[:SENSe]:MCPower:CARRier:LIST:BANDwidth|BWIDth:[INTegration]?
Displays menu keys that enable you to configure the offsets and limits for the MCP measurement.
There is no equivalent remote command.
5.19.4.1 Offset
Enables you to select the offset the menu keys will affect. Press Offset until the letter of the desired offset (A, B, or C) is underlined.
There is no equivalent remote command.
5.19.4.2 Offset Freq
Sets the frequency difference between the center of the uppermost carrier and the center of the upper element of the offset pair, as well as the frequency difference between the center of the lower element of the offset pair and the center of the lowermost carrier. If you set the frequency of any offset to 0 Hz, the offset is turned off and not included in the displayed results.
State Saved: Saved in instrument state.

Meas Setup (Multi-Carrier Power—MCP)
[:SENSe]:MCPower:OFFSet:LIST:[FREQuency],,
[:SENSe]:MCPower:OFFSet:LIST:[FREQuency]?
5.19.4.3 Offset Integ BW
Sets the bandwidth over which the power is integrated for the selected offset (refer to “Offset” above).
Key Notes: If RRC Filter is set to On, the actual integration bandwidth used is the displayed integration bandwidth multiplied by (1 + Filter Alpha).
State Saved: Saved in instrument state.
Step Key Increment: If CF Step(Auto) is selected: span/10 If CF Step(Man) is selected: CF Step
[:SENSe]:MCPower:OFFSet:LIST:BANDwidth|BWIDth:[INTegration],,
[:SENSe]:MCPower:OFFSet:LIST:BANDwidth|BWIDth:[INTegration],,
5.19.4.4 Upper Offset Limit
Sets the limit for the selected offset (refer to “Offset” above) to the right of the carriers.

Meas Setup (Multi-Carrier Power—MCP)
State Saved: Saved in instrument state.
:CALCulate:MCPower:OFFSet:LIST:LIMit:POSitive[:UPPer]:DATA ,,
:CALCulate:MCPower:OFFSet:LIST:LIMit:POSitive[:UPPer]:DATA?
5.19.4.5 Lower Offset Limit
Sets the limit for the selected offset (refer to “Offset” above) to the left of the carriers.
State Saved: Saved in instrument state.
:CALCulate:MCPower:OFFSet:LIST:LIMit:NEGative[:UPPer]:DATA ,,
:CALCulate:MCPower:OFFSet:LIST:LIMit:NEGative[:UPPer]:DATA?

Meas Setup (Multi-Carrier Power—MCP)
Press Carrier Result to select the result you wish to display on the last line of the carrier power results list except when:
Key Notes: This key is only available when Meas Control, Measure (Single) is selected.
State Saved: Not saved.
Range: 1 to number of carriers.
History Added with firmware revision A.03.00
There is no equivalent remote command.
Sets the input attenuator to optimize the robustness of the measurement, which is its freedom from errors due to input compression.
State Saved: Not saved.
[:SENSe]:POWer[:RF]:RANGe:AUTO ONCE
Enables you to set the measurement method to either the integration bandwidth method (IBW) or the resolution bandwidth method (RBW). The resolution bandwidth method is

Meas Setup (Multi-Carrier Power—MCP)
most useful for measuring cdmaOne and cdma2000 signals; the integration bandwidth medthod is preferred for other signals.
Key Notes: When Method is set to RBW, neither Noise Correction nor RRC Filter are available.
State Saved: Saved in instrument state.
History: Added with firmware revision A.02.00.
[:SENSe]:MCPower:METHod IBW|RBW
[:SENSe]:MCPower:METHod?
5.19.8 Power Ref
Enables you to set the multi-carrier power reference to automatic or manual. When set to automatic, the carrier power result reflects the measured power value in the selected reference carrier (Meas Setup, Carrier Setup, Ref Carrier). When set to manual, the result is referenced to the last measured value, or you may specify the reference for the multi-carrier power measurement. Relative values are displayed, referenced to the “Power Reference” value.
State Saved: Saved in instrument state.
Factory Preset: Off/Measured power in the reference carrier.
History: Added with firmware revision A.02.00.
[:SENSe]:MCPower:CARRier:AUTO[:STATe] OFF|ON|0|1
[:SENSe]:MCPower:CARRier:AUTO[:STATe]?
[:SENSe]:MCPower:CARRier[:POWer]
[:SENSe]:MCPower:CARRier[:POWer]?

Meas Setup (Multi-Carrier Power—MCP)
MCP:CARR -100
MCP:CARR?
5.19.9 Limit Test
Pressing Limit Test turns the testing of the limit line on or off. When Limit Test is set to On, each offset is compared to its upper and lower offset limit. In those cases where the power exceeds the limit, a red “F” is placed next to the dBc result to indicate a failure. If there are any failures, the pass/fail indicator area show “FAIL” in red; if there are none, it shows “PASS” in green. Any offsets that are in the off state (see “Offset Freq” on page 438) are not measured and their results will not be displayed on screen.
State Saved: Saved in instrument state.

Meas Setup (Multi-Carrier Power—MCP)
[:SENSe]:MCPower:LIMit[:STATe] OFF|ON|0|1
[:SENSe]:MCPower:LIMit[:STATe]?
5.19.10 RRC Filter
Pressing RRC Filter turns the Root Raised Cosine filter on or off for the carriers and all adjacent channels. The RRC filter is the type specified in the 3GPP W-CDMA standards, with roll-off (α) = Filter Alpha parameter (defined below).T is set to 1/Carrier Integ BW (for carriers) or 1/Offset Integ BW (for Offsets).
Key Notes: This key is not available when Meas Setup, Method (RBW) is selected.
State Saved: Saved in instrument state.
[:SENSe]:MCPower:FILTer[:RRC][:STATe] OFF|ON|0|1
[:SENSe]:MCPower:FILTer[:RRC][:STATe]?
5.19.11 Filter Alpha
Press Filter to input the alpha value for the RRC Filter. This parameter is only available when 3GPP W-CDMA or None has been selected as the radio standard from the Mode Setup menu.
State Saved: Saved in instrument state.
Factory Preset: 0.22 when W-CDMA is selected, otherwise Off.
History: Added with firmware revision A.03.00.

Meas Setup (Multi-Carrier Power—MCP)
[:SENSe]:MCPower:FILTer[:RRC]:ALPHA
[:SENSe]:MCPower:FILTer[:RRC]:ALPHA?
5.19.12 Noise Correction
Pressing Noise Correction turns noise correction on or off. When you set Noise Correction to On, a calibration of the noise floor is performed and used to correct for analyzer noise floor contribution to measurement levels, increasing dynamic range.
Key Notes: This key is not available when Meas Setup, Method (RBW) is selected or when signal tracking is on.
State Saved: Saved in instrument state.
History: Added with firmware revision A.02.00.
[:SENSe]:MCPower:CORRection:NOISe[:AUTO] OFF|ON|0|1
[:SENSe]:MCPower:CORRection:NOISe[:AUTO]?
Remote Command Notes: The noise correction feature is not available when the measurement method is RBW (Meas Setup, Method (RBW) or when signal tracking is on.

Meas Setup (Multi-Carrier Power—MCP)

Trace/View (Multi-Carrier Power Measurement)
If Multi-Carrier Power has been selected in the Measure menu of the Spectrum Analysis mode, this key displays the appropriate Trace/View menu for the multi-carrier power measurement.
Displays menu keys that enable you to set how trace information is stored and displayed.
There is no equivalent remote command for this function.
Pressing Spectrum selects the spectral display of the measurement.
There is no remote command for this function.
Pressing Combined selects the measurement to be displayed as a bar graph and spectrum.
There is no remote command for this function.

Trace/View (Multi-Carrier Power Measurement)
Pressing Combined View Units selects the units (dBc or dBm) for the floating numeric displays when Combined is selected.
There is no remote command for this function.
One-button measurements do not change this function. See “Trace” on page 284 for more information.
5.20.4.1 Trace (1 2 3)
Determines which trace the menu keys will affect. Press Trace 1 2 3 until the number of the desired trace is underlined.
:TRACe[1]|2|3:MODE WRITe|MAXHold|MINHold|VIEW|BLANk
WRITe = Clear Write
MAXHold = Max Hold
MINHold = Min Hold
:TRACe[1]|2|3:MODE?
TRAC:MODE MINH
TRAC:MODE VIEW
TRAC:MODE BLANk
VIEW = View · BLANk = Blank

Trace/View (Multi-Carrier Power Measurement)

Trace/View (Multi-Carrier Power Measurement)
5.20.4.2 Clear Write
See “Clear Write” on page 285 for more information.
5.20.4.3 Max Hold
See “Max Hold” on page 285 for more information.
5.20.4.4 Min Hold
See “Min Hold” on page 285 for more information.
5.20.4.5 View
See “View” on page 286 for more information.
5.20.4.6 Blank
See “Blank” on page 286 for more information.

Meas Setup (Occupied Bandwidth—OBW)
When Occupied BW has been selected in the Measure menu of the Spectrum Analysis Mode, this key displays the appropriate measurement setup menu.
The Occupied Bandwidth measurement integrates the power of the displayed spectrum and puts markers at the frequencies between which a selected percentage of the power is contained. The measurement defaults to 99% of the occupied bandwidth power. The power-bandwidth routine first computes the combined power of all signal responses contained in the trace. For 99% occupied power bandwidth, markers are placed at the frequencies on either side of 99% of the power. 1% of the power is evenly distributed outside the markers. The difference between the marker frequencies is the 99% power bandwidth and is the value displayed.
The occupied bandwidth function also indicates the difference between the analyzer center frequency and the center frequency of the channel, referred to as “Transmit Freq Error.” The measurement can be made in single or continuous sweep mode. The center frequency, reference level, and channel spacing must be set by the user.
Press Avg Number (On) to specify the number of measurement averages used when calculating the measurement result. The average is displayed at the end of each sweep. Press Avg Number (Off) to disable measurement averaging.
State Saved: Saved in instrument state.
[:SENSe]:OBW:AVERage:COUNt
[:SENSe]:OBW:AVERage:COUNt?
[:SENSe]:OBW:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:OBW:AVERage[:STATe]?

Meas Setup (Occupied Bandwidth—OBW)
OBW:AVER?
5.21.2 Avg Mode
Enables you to select the type of termination control used for the averaging function (Exp or Repeat). This determines the averaging action after the specified number of measurements (average count) is reached.
State Saved: Saved in instrument state.
[:SENSe]:OBW:AVERage:TCONtrol EXPonential|REPeat
[:SENSe]:OBW:AVERage:TCONtrol?
5.21.3 Max Hold
Enables you to turn maximum hold trace feature On or Off for the measurement. Maximum hold displays and holds the maximum responses of a signal.
State Saved: Saved in instrument state.
[:SENSe]:OBW:MAXHold OFF|ON|0|1
[:SENSe]:OBW:MAXHold?

Meas Setup (Occupied Bandwidth—OBW)

Meas Setup (Occupied Bandwidth—OBW)
Enables you to change the percentage of signal power used when determining the occupied bandwidth.
State Saved: Saved in instrument state.
[:SENSe]:OBW:PERCent
[:SENSe]:OBW:PERCent?
5.21.5 OBW Span
Enables you to specify the range of integration used in calculating the total power from which the percent occupied bandwidth is then calculated. The analyzer span will be set to the same value as the OBW Span for the measurement. OBW Span should be set to approximately 2 times the expected occupied bandwidth result.
Factory Preset: 3 MHz, or as defined by the radio standard selected.
Range:
[:SENSe]:OBW:FREQuency:SPAN
[:SENSe]:OBW:FREQuency:SPAN?

Meas Setup (Occupied Bandwidth—OBW)
Enables you to specify the power level used to determine the emission bandwidth as the number of dB down from the highest signal point (P1), within the occupied bandwidth span. Frequencies f1 and f2 are determined as the furthest frequencies xdB below and above P1, respectively. The emission bandwidth is then calculated as f2 −f1 as shown in
Figure 5-1.
NOTE The asterisk next to the x dB bandwidth value in Figure 5-1 indicates the results may not have been determined with optimal analyzer settings. If emission bandwidth is your primary interest, select Meas Setup, Max Hold. Then change detector mode to peak by pressing Det/Demod, Detector, Peak. Acquiring peak data ensures accuracy of this measurement.
Figure 5-1 Occupied Bandwidth Measurement Results

f2 f1fff1
State Saved: Saved in instrument state.

Meas Setup (Occupied Bandwidth—OBW)
[:SENSe]:OBWidth:XDB
[:SENSe]:OBWidth:XDB?
5.21.7 Optimize Ref Level
Sets the input attenuator to optimize the robustness of the measurement, which is its freedom from errors due to input compression.
State Saved: Saved in instrument state.
[:SENSe]:POWer[:RF]:RANGe:AUTO ONCE

Meas Setup (Spectrum Emissions Mask—SEM)
When the spectrum emissions mask measurement has been selected in the Measure menu of the Spectrum Analysis Mode, this key displays the appropriate measurement setup menu.
Spectrum Emissions Mask (SEM) measurement includes the in-band and out-of-band spurious emissions. As it applies to W-CDMA (3GPP), this is the power contained in a specified frequency bandwidth at certain offsets relative to the total carrier power. It may also be expressed as a ratio of power spectral densities between the carrier and the specified offset frequency band.
Added with firmware revision A.02.00
Press Avg Number (On) to specify the number of measurement averages used when calculating the measurement result. The average will be displayed at the end of each sweep.
State Saved: Saved in instrument state.
[:SENSe]:SEMask:AVERage:COUNt
[:SENSe]:SEMask:AVERage:COUNt?
[:SENSe]:SEMask:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:SEMask:AVERage[:STATe]?

Meas Setup (Spectrum Emissions Mask—SEM)
Displays a menu where you can select a measurement reference type, Total Pwr Ref or PSD Ref.
[:SENSE]:SEMask:TYPE TPRef | PSDRef
[:SENSE]:SEMask:TYPE?
5.22.2.1 Total Pwr Ref
When Total Pwr Ref is selected, the power in the carrier is used as the reference in computing the relative power values for the offsets.
Readback: Reads back “Total Pwr Ref” to the Meas Type key.
5.22.2.2 PSD Ref
When PSD Ref is selected, the mean power spectral density of the carrier is used in computing the relative power spectral density values for the offsets.
Readback: Reads back “PSD Ref” to the Meas Type key.
Accesses the menu of keys that enable you to change the reference channel settings.

Meas Setup (Spectrum Emissions Mask—SEM)
5.22.3.1 Chan Integ BW
Specifies the integration bandwidth used in calculating the power in the main channel.
[:SENSe]:SEMask:BANDwidth[n]|BWIDth[n]:INTegration
[:SENSe]:SEMask:BANDwidth[n]|BWIDth[n]:INTegration?
SEM:BWID[n]:INT 4 MHz
SEM:BAND[n]:INT?
SEM:BWID[n]:INT?
5.22.3.2 Chan Span
Specifies the span used in measuring the power in the main channel.
Knob Increment: 1% of the set span.
[:SENSe]:SEMask:FREQuency:SPAN
[:SENSe]:SEMask:FREQuency:SPAN?
5.22.3.3 Sweep Time
Specifies the sweep time used in measuring the power in the main channel.

Meas Setup (Spectrum Emissions Mask—SEM)
[:SENSe]:SEMask:SWEeptime
[:SENSe]:SEMask:SWEeptime?
[:SENSe]:SEMask:SWEeptime:AUTO OFF|ON|0|1
[:SENSe]:SEMask:SWEeptime:AUTO?
Example: SEM:SWE 4 s SEM:SWE? SEM:SWE:AUTO 1 SEM:SWE:AUTO?
5.22.3.4 Res BW
Specifies the resolution bandwidth used in measuring and the power in the main channel.
Range: Full Range provided by Base Instrument Hardware.
[:SENSe]:SEMask:BANDwidth|BWIDth[:RESolution]
[:SENSe]:SEMask:BANDwidth|BWIDth[:RESolution]?
[:SENSe]:SEMask:BANDwidth|BWIDth[:RESolution]:AUTO OFF|ON|0|1
[:SENSe]:SEMask:BANDwidth|BWIDth[:RESolution]:AUTO?
Remote Command Notes: If an unavailable bandwidth is entered, the closest available bandwidth is used. Refer to “Res BW” on page 83.
Example: SEM:BAND 4 MHz SEM:BWID 4 MHz SEM:BAND? SEM:BWID?

Meas Setup (Spectrum Emissions Mask—SEM)
5.22.3.5 Total Pwr Ref or PSD Ref
Displays the type of power measurement reference selected using the Meas Type key (see “Meas Type” on page 458).
Total Pwr Ref is the power in the carrier that is used as the reference in computing the relative power values for the offsets. When Man is selected, this can be set by the user. When Auto is selected, this is the measured power in the reference carrier.
PSD Ref is the mean power spectral density in the carrier that is used as the reference in computing the relative power spectral density values for the offsets. When Man is selected, this can be set by the user, when Auto is selected, this is the measured mean power spectral density in the reference carrier.
Range: –200.0 dBm to 200.0 dBm.

Meas Setup (Spectrum Emissions Mask—SEM)
[:SENSe]:SEMask:CARRier[:POWer]
[:SENSe]:SEMask:CARRier[:POWer]?
[:SENSe]:SEMask:CARRier:AUTO[:STATe] OFF|ON|0|1
[:SENSe]:SEMask:CARRier:AUTO[:STATe]?
SEM:CARR:AUTO 1
SEM:CARR:AUTO?
5.22.4 Offset/Limits
Displays the menus where you can change the following parameters for offset frequency settings and pass/fail tests: Offset, Start Freq, Stop Freq, Sweep Time, Res BW, and Limits. If Spectrum Segment is set to Offset. Refer to the tables below for default settings of particular radio standards of interest.
Table 5-2 Offsets & Limits Defaulted for W-CDMA (3GPP) Base Station Measurementsa
Offset Start Freq (MHz)
A, On 2.515 2.715 30.00 −14.00 −14.00 Couple −30.00 −30.00 1
B, On 2.715 3.515 30.00 −14.00 −26.00 Man −30.00 −30.00 1
C, On 3.515 4.000 30.00 −26.00 −26.00 Couple −30.00 −30.00 1
D, On 4.000 8.000 1000.0 −13.00 −13.00 Couple −30.00 −30.00 1
E, On 8.000 12.500 1000.0 −13.00 −13.00 Couple −30.00 −30.00 1
a. Fail Mask is set to Absolute and Rel Stop is set to Couple.

Meas Setup (Spectrum Emissions Mask—SEM)
Table 5-3 Offsets & Limits Defaulted for W-CDMA (3GPP) Mobile Station Measurementsa
Offset Start Freq (MHz)
A, On 2.515 3.485 30.00 −71.07 −71.07 −35.23 −49.78 Man 1
B, On 4.000 7.500 1000.0 −55.84 −55.84 −35.50 −39.00 Man 1
C, On 7.500 8.500 1000.0 −55.84 −55.84 −39.00 −49.00 Man 1
D, On 8.500 12.000 1000.0 −55.84 −55.84 −49.00 −49.00 Couple 1
E, Off 12.500 15.000 1000.0 −55.84 −55.84 −49.00 −49.00 Couple 1
a. Abs Stop is set to Couple and Fail Mask is set to Abs AND Rel.
Table 5-4 Offsets & Limits Defaulted for 802.11a and HiperLAN/2 Radio Standards a
Offset Start Freq (MHz)
A, On 9.00 11.00 16.00 −4.00 Man 0.00 −20.00 Man 1
B, On 11.00 20.00 −4.00 −12.00 Man −20.00 −28.00 Man 1
C, On 20.00 30.00 −12.00 −24.00 Man −28.00 −40.00 Man 1
D, On 30.00 50.00 −24.00 −24.00 Couple −40.00 −40.00 Couple 1
E, Off 50.00 100.00 −24.00 −24.00 Couple −40.00 −40.00 Couple 1
a. Sweep Time is set to 100 mseconds, Res BW is set to 100 kHz, and Fail Mask is
set to Rel.
Table 5-5 Offsets & Limits Defaulted for 802.11b Radio Standarda
Offset Start Freq (MHz)
A, On 11.00 22.00 −10.00 −10.00 −30.00 −30.00 1
B, On 22.00 50.00 −30.00 −30.00 −50.00 −50.00 1
C, Off 50.00 70.00 −30.00 −30.00 −50.00 −50.00 1
D, Off 70.00 90.00 −30.00 −30.00 −50.00 −50.00 1
E, Off 90.00 100.00 −30.00 −30.00 −50.00 −50.00 1
a. Sweep Time is set to 100 mseconds, Res BW is set to 100 kHz, Fail Mask
is set to Rel, and Abs Stop and Rel Stop are both set to Couple

Meas Setup (Spectrum Emissions Mask—SEM)
5.22.4.1 Offset
Selects the offset pairs (upper and lower) that the menu keys affect, and displays the memory selection menu from A to E (where you can store up to 5 sets of values for Start Freq, Stop Freq, Sweep Time, Res BW, Meas BW, Abs Start, Abs Stop, and so forth). Press Offset until the letter of the desired offset (A, B, C, D, or E) is underlined. Only one selection at a time is shown on this key label.
State Saved: Saved in instrument state.
There is no remote command for this function.
5.22.4.2 Start Freq
Specifies the inner limit (frequency closest to the carrier) for both segments of the specified offset pair. When Start Freq (Off) is selected, the offset pair is not measured.
State Saved: Saved in instrument state.
tables for “Offset/Limits” on page 462 for this value.
[:SENSe]:SEMask:OFFSet[n]:LIST:FREQuency:STARt , ... [] (up to five values)
[:SENSe]:SEMask:OFFSet[n]:LIST:FREQuency:STARt?
[:SENSe]:SEMask:OFFSet[n]:LIST:STATe OFF|ON|0|
[:SENSe]:SEMask:OFFSet[n]:LIST:STATe?
Remote Command Notes: Comma separated list of up to 5 values. n = 1 for BTS and n = 2 for MS. Default is BTS.
Offsets that are turned off, for the currently selected offset will return a -999.0 when queried.

Meas Setup (Spectrum Emissions Mask—SEM)
SEM:OFFS:LIST:STAT?
5.22.4.3 Stop Freq
Specifies the outer limit (frequency furthest from the carrier) for both segments of the specified offset pair. The lower range is limited to the setting of Start Freq.
State Saved: Saved in instrument state.
tables for “Offset/Limits” on page 462 for this value.
[:SENSe]:SEMask:OFFSet[n]:LIST:FREQuency:STOP ... [] (up to five values)
[:SENSe]:SEMask:OFFSet[n]:LIST:FREQuency:STOP?
Remote Command Notes: Comma separated list of up to 5 values. n = 1 for BTS and n = 2 for MS. Default is BTS.
5.22.4.4 Sweep Time
Specifies the sweep time for the currently selected offset, and toggles this function between Auto and Manual for each offset.
State Saved: Saved in instrument state.
tables for “Offset/Limits” on page 462 for this value.
[:SENSe]:SEMask:OFFSet[n]:LIST:SWEeptime ... [] (up to five

Meas Setup (Spectrum Emissions Mask—SEM)
[:SENSe]:SEMask:OFFSet[n]:LIST:SWEeptime?
[:SENSe]:SEMask:OFFSet[n]:LIST:SWEeptime:AUTO OFF|ON|0|1
[:SENSe]:SEMask:OFFSet[n]:LIST:SWEeptime:AUTO?
Remote Command Notes: Comma separated list of up to 5 values. n = 1 for BTS and n = 2 for MS. Default is BTS.
SEM:OFFS:LIST:SWE 4 ms
5.22.4.5 Res BW
Specifies the resolution bandwidth used in measuring the offset pair. When set to Auto, the Res BW from the default tables for “Offset/Limits” on page 462 are used. When set to Man, the range of settings is the range of the available Res BW of the analyzer, except the maximum is further limited to not exceed (Stop Freq - Start Freq).
State Saved: Saved in instrument state.
tables for “Offset/Limits” on page 462 for this value.
[:SENSe]:SEMask:OFFSet[n]:LIST:BANDwidth|BWIDth[:RESolution] ...[] (up to five values)
[:SENSe]:SEMask:OFFSet[n]:LIST:BANDwidth|BWIDth[:RESolution]?
[:SENSe]:SEMask:OFFSet[n]:LIST:BANDwidth|BWIDth[:RESolution]:AUTO OFF|ON|0|1
[:SENSe]:SEMask:OFFSet[n]:LIST:BANDwidth|BWIDth[:RESolution]:AUTO?
Remote Command Notes: Comma separated list of up to5 values. n = 1 for BTS and n = 2 for MS. Default is BTS (1).
You may only enter valid Res BW filter frequencies. Refer to “Res BW” on page 83.
values)

Meas Setup (Spectrum Emissions Mask—SEM)
SEM:OFFS:LIST:BAND? or SEM:OFFS:LIST:BWID?
SEM:OFFS:LIST:BAND:AUTO 0 or SEM:OFFS:LIST:BWID:AUTO 0
SEM:OFFS:LIST:BAND:AUTO? or SEM:OFFS:LIST:BWID:AUTO?

Meas Setup (Spectrum Emissions Mask—SEM)
5.22.4.6 Meas BW
Enables you to specify the bandwidth to use when measuring the offset. When using the front panel, only the currently selected offset is affected. All the offsets are affected when using the remote command. This is the multiplier applied to the resolution bandwidth to determine the bandwidth to integrate when calculating the power.
Key Notes: When setting Meas BW > 1, dynamic range is increased, however measurement time is also increased since an increased number of sweep points are required to make the measurement. For Meas BW > 1, consider this equation for trace points:
Trace points = (Offset Stop Freq - Offset Start Freq)/ (Meas BW/100)
When Meas BW = 1, the number of trace points are set at the value specified upon entering the measurement.
Dependencies/ Couplings: This parameter must adhere to the rule: (N x Res BW) ≤ (Stop Freq of the offset - Start freq of the offset), where N is Meas BW value. If the Res BW is changed, Meas BW will change to ensure this rule.
State Saved: Saved in instrument state.
[:SENSe]:SEMask:OFFSet[n]:LIST:BANDwidth|BWIDth:IMULti
[:SENSe]:SEMask:OFFSet[n]:LIST:BANDwidth|BWIDth:IMULti?
[:SENSe]:SEMask:OFFSet[n]:LIST:BANDwidth|BWIDth[:RESolution]:AUTO OFF|ON|0|1
[:SENSe]:SEMask:OFFSet[n]:LIST:BANDwidth|BWIDth[:RESolution]:AUTO?
Remote Command Notes: Comma separated list of up to 5 values. n = 1 for BTS and n = 2 for MS. Default is BTS (1).
n/a

Meas Setup (Spectrum Emissions Mask—SEM)
SEM:OFFS:LIST:BAND:IMUL? or SEM:OFFS:LIST:BWID:IMUL?
SEM:OFFS:LIST:BAND:AUTO 0 or SEM:OFFS:LIST:BWID:AUTO 0
SEM:OFFS:LIST:BAND:AUTO? or SEM:OFFS:LIST:BWID:AUTO?
5.22.4.7 Abs Start
Enables you to enter an absolute level limit at Start Freq for the currently selected offset ranging from −200.00 to +50.00 dBm with 0.01 dB resolution.
State Saved: Saved in instrument state.
tables for “Offset/Limits” on page 462 for this value.
[:SENSe]:SEMask:OFFSet[n]:LIST:STARt:ABSolute ,...[] (up to five values)
[:SENSe]:SEMask:OFFSet[n]:LIST:STARt:ABSolute?
Remote Command Notes: Comma separated list of up to 5 values. n = 1 for BTS and n = 2 for MS. Default is BTS (1).
5.22.4.8 Abs Stop
Enables you to enter an absolute level limit at Stop Freq for the currently selected offset ranging from −200.00 to +50.00 dBm with 0.01 dB resolution, and to toggle this function between Couple and Man. If set to Couple, this is coupled to Abs Start to make a flat limit line. If set to Man, Abs Start and Abs Stop you can enter different values to make a sloped limit line.
State Saved: Saved in instrument state.
tables for “Offset/Limits” on page 462 for this value.

Meas Setup (Spectrum Emissions Mask—SEM)
[:SENSe]:SEMask:OFFSet[n]:LIST:STOP:ABSolute ,...[] (up to five values)
[:SENSe]:SEMask:OFFSet[n]:LIST:STOP:ABSolute?
[:SENSe]:SEMask:OFFSet[n]:LIST:STOP:ABSolute:COUPle OFF|ON|0|1
[:SENSe]:SEMask:OFFSet[n]:LIST:STOP:ABSolute:COUPle?

Meas Setup (Spectrum Emissions Mask—SEM)
Remote Command Notes: Comma separated list of up to 5 values. n = 1 for BTS and n = 2 for MS. Default is BTS (1).
SEM:OFFS:LIST:STOP:ABS:COUP 0
SEM:OFFS:LIST:STOP:ABS:COUP?
5.22.4.9 Rel Start
Enables you to enter a relative level limit at Start Freq ranging from −150.00 to +50.00 dBc with 0.01 dB resolution.
State Saved: Saved in instrument state.
tables for “Offset/Limits” on page 462 for this value.
[:SENSe]:SEMask:OFFSet[n]:LIST:STARt:RCARrier ,... [] (up to five values)
[:SENSe]:SEMask:OFFSet[n]:LIST:STARt:RCARrier?
Remote Command Notes: Comma separated list of 5 values. n = 1 for BTS and n = 2 for MS. Default is BTS (1).
5.22.4.10 Rel Stop
Enables you to enter a relative level limit at Stop Freq ranging from −150.00 to +50.00 dBc with 0.01 dB resolution, and to toggle this function between Couple and Man. If set to Couple, this is coupled to Rel Start to make a flat limit line. If set to Man, Rel Start and Rel Stop you can enter different values to make a sloped limit line.
State Saved: Saved in instrument state.
tables for “Offset/Limits” on page 462 for this value.

Meas Setup (Spectrum Emissions Mask—SEM)
[:SENSe]:SEMask:OFFSet[n]:LIST:STOP:RCARrier ,[]...
[:SENSe]:SEMask:OFFSet[n]:LIST:STOP:RCARrier?
[:SENSe]:SEMask:OFFSet[n]:LIST:STOP:RCARrier:COUPle OFF|ON|0|1
[:SENSe]:SEMask:OFFSet[n]:LIST:STOP:RCARrier:COUPle?
Remote Command Notes: Comma separated list of 5 values. n = 1 for BTS and n = 2 for MS. Default is BTS (1).
SEM:OFFS:LIST:STOP:RCAR:COUP 0
SEM:OFFS:LIST:STOP:RCAR:COUP?
5.22.4.11 Fail Mask
Displays the menu to select one of the following logic keys for fail conditions between the measurement results and the test limits: Absolute, Relative, Abs AND Rel, and Abs OR Rel.
Absolute - Fail is shown if one of the absolute spectrum emission mask measurement results is larger than the limit between Abs Start and Abs Stop, inclusive. This is the default selection for each offset for BTS measurements.
Relative - Fail is shown if one of the relative spectrum emission mask measurement results is larger than the limit between Rel Start and Rel Stop, inclusive.
Abs AND Rel - Fail is shown if one of the absolute spectrum emission mask measurement results is larger than the limit between Abs Start and Abs Stop, inclusive AND one of the relative spectrum emission mask measurement results is larger than the limit for Rel Start and Rel Stop. This is the default for MS measurements.
Abs OR Rel - Fail is shown if one of the absolute spectrum emission mask measurement results is larger than the limit between Abs Start and Abs Stop, inclusive OR one of the relative spectrum emission mask measurement results is larger than the limit for Rel Start and Rel Stop.
State Saved: Saved in instrument state.

Meas Setup (Spectrum Emissions Mask—SEM)
tables for “Offset/Limits” on page 462 for this value.

Meas Setup (Spectrum Emissions Mask—SEM)
[:SENSe]:SEMask:OFFSet[n]:LIST:TEST ABSolute|AND|OR|RELative,ABSolute|AND|OR|RELative,ABSolute|AND|OR|RELative ,ABSolute|AND|OR|RELative
[:SENSe]:SEMask:OFFSet[n]:LIST:TEST?
Remote Command Notes: Comma separated list of 5 values. n = 1 for BTS and n = 2 for MS. Default is BTS (1).
Example: SEM:OFFS:LIST:TEST ABS,REL,ABS AND REL,ABS OR REL SEM:OFFS:LIST:TEST?
5.22.4.11.1 Absolute
Set the limit test to show Fail if one of the absolute spectrum emission mask measurement results is larger than the limit for Abs Start and/or Abs Stop. This is the default selection for each offset
State Saved: Saved in instrument state.
tables for “Offset/Limits” on page 462 for this value.
Refer to “Fail Mask” on page 472.
5.22.4.11.2 Relative
Sets the limit test to show Fail if one of the relative spectrum emission mask measurement results is larger than the limit for Rel Start and/or Rel Stop.
State Saved: Saved in instrument state.
tables for “Offset/Limits” on page 462 for this value.
Refer to “Fail Mask” on page 472.
5.22.4.11.3 Abs AND Rel
Set the limit test to show Fail if one of the absolute spectrum emission mask measurement results is larger than the limit for Abs Start and Abs Stop AND one of the relative spectrum emission mask measurement results is larger than the limit for Rel Start and Rel Stop.

Meas Setup (Spectrum Emissions Mask—SEM)
State Saved: Saved in instrument state.
tables for “Offset/Limits” on page 462 for this value.
Refer to “Fail Mask” on page 472.
5.22.4.11.4 Abs OR Rel
Set the limit test to show Fail if one of the absolute spectrum emission mask measurement results is larger than the limit for Abs Start and Abs Stop OR one of the relative spectrum emission mask measurement results is larger than the limit for Rel Start and Rel Stop.
State Saved: Saved in instrument state.
tables for “Offset/Limits” on page 462 for this value.
Refer to “Fail Mask” on page 472.
Sets the input attenuator to optimize the robustness of the measurement, which is its freedom from errors due to input compression.
State Saved: Saved in instrument state.
[:SENSe]:POWer[:RF]:RANGe:AUTO ONCE
Pressing RRC Filter turns the Root Raised Cosine filter on or off. This filter is the type

Meas Setup (Spectrum Emissions Mask—SEM)
specified in the 3GPP W-CDMA standards. This parameter is only available when 3GPP W-CDMA is selected as the Radio Std from the Mode Setup menu. If 3GPP W-CDMA is selected, the roll-off value (alpha) for the filter will be initially set to 0.22 and T will be 260 ns. The roll-off value can be changed using Filter Alpha.
State Saved: Saved in instrument state.
History: Added with firmware revision A.03.00.
[:SENSe]:SEMask:FILTer[:RRC][:STATe] OFF|ON|0|1
[:SENSe]:SEMask:FILTer[:RRC][:STATe]?
5.22.7 Filter Alpha
Press RRC Filter to input the alpha value for the RRC Filter. This parameter is only available when 3GPP W-CDMA is selected as the Radio Std. from the Mode Setup menu.
State Saved: Saved in instrument state.
Factory Preset: 0.22 when W-CDMA is selected, otherwise Off.
History: Added with firmware revision A.03.00.
[:SENSe]:SEMask:FILTer[:RRC]:ALPHA
[:SENSe]:SEMask:FILTer[:RRC]:ALPHA?

Trace/View (Spectrum Emissions Mask)
If Spectrum Emission Mask has been selected in the Measure menu of the Spectrum Analysis mode, this key displays the appropriate Trace/View menu for the spectrum emission mask (SEM).
Displays menu keys that enable you to set how trace information is stored and displayed.
There is no remote command for this function.
Press Abs Pwr & Freq to view the measurement results of Spectrum (Ref: Total Pwr, Seg: Offset) measurements in the graph window as absolute peak power and frequency. The absolute peak power levels, in dBm and those corresponding offset frequency ranges on both sides of the reference channel are displayed in the text window.
Saved State: Saved in instrument state.
There is no remote command for this function.
Press Rel Pwr & Freq to view the measurement results of Spectrum (Ref: Total Pwr, Seg: Offset) measurements in the graph window as power relative to the carrier power and frequency. The relative power levels in dBc and those corresponding offset frequency ranges on both sides of the reference channel are displayed in the text window.
Saved State: Saved in instrument state.
There is no remote command for this function.

Trace/View (Spectrum Emissions Mask)
Press Integrated Power to view the measurement results of Spectrum (Ref: Total Pwr, Seg: Offset) measurements in the graph window as integrated absolute and relative peak power and frequency. The absolute and relative peak power levels integrated throughout the bandwidths between the start and stop frequencies and those corresponding offset frequency ranges on both sides of the reference channel are displayed in the text window.
Saved State: Saved in instrument state.
There is no remote command for this function.

Display (Spectrum Emissions Mask—SEM)
Displays menu keys that enable you to configure parameters for the CCDF measurement.
Expands the measurement window horizontally over the entire instrument display. Pressing any other key that results in a new menu cancels the full screen function.
State Saved: Not saved in state.
Allows you to toggle the limit lines display function for spectrum emission mask measurements between On and Off. If set to On, the absolute limit lines and the relative limit lines are shown on the spectrum emission mask measurement display.
State Saved: Saved in instrument state.
Displays menu keys that enable you to turn the graticule and annotation on or off.
State Saved: Saved in instrument state.

Display (Spectrum Emissions Mask—SEM)
5.24.3.1 Graticule (On Off)
Pressing Graticule turns the display graticules On or Off.
5.24.3.2 Annotation (On Off)
Turns the screen annotation on or off, however, menu key annotation will remain on the display. The screen annotation may not be required for prints or during remote operation.

SPAN X Scale (Spectrum Emissions Mask—SEM)
Activates the Span function for the SEM measurement and displays the menu of span functions.
See the Scale/Div, Ref Value, and Ref Position commands below.
Allows you to enter a numeric value to change the horizontal display sensitivity by setting the frequency units per horizontal graticule division. To determine full display frequency span, multiply the frequency span per horizontal graticule division by 10. This function is only available when you select Spectrum Emissions Mask from the Measurement menu.
Press Ref Value to set the frequency represented by the selected horizontal graticule line on the display (the reference). The current value is indicated by Ref Value in the upper left corner of the display. You can change the reference value using the step keys, the knob, or the numeric keypad. You can set the reference line location using the Ref Position setting. This function is only available when you select Spectrum Emissions Mask from the Measurement menu.
Factory Preset: Center frequency value in Factory Preset state.
Press Ref Position to position the X-Scale reference to the extreme left line, center line, or

SPAN X Scale (Spectrum Emissions Mask—SEM)
extreme right line of the display graticule.

Marker (Spectrum Emissions Mask—SEM)
Displays the Marker control menu for the SEM measurement. If no markers are active, Marker selects marker 1, sets it to Normal and places it to the right of the display. There are two control modes for the markers:
State Saved: The control mode for each marker, as well as the position of each marker, is saved in instrument state.
Factory Preset: All Off.
There is no equivalent remote command.
Selects one of the four possible markers. Once a marker is selected, it can be set to any of the control modes Normal or Off.
State Saved: The number of the selected marker is saved in instrument state.
:CALCulate:SEMask:MARKer[1]|2|3|4:STATe OFF|ON|0|1
:CALCulate:SEMask:MARKer[1]|2|3|4:STATe?
Remote Command Notes: Sets or queries the state of a marker. Setting a marker to state ON or 1 selects that marker. Setting a marker which is OFF to state ON or 1 puts it in Normal mode and places it on the screen to the right of the displayed measurement graph. Setting a marker to state OFF or 0 selects that marker and turns it off. The response to the query will be 0 if OFF, 1 if ON.
Example: CALC:SEM:MARK2:STAT ON selects marker 2.
CALC:SEM:MARK:STAT ON will not modify a marker that is already on.

Marker (Spectrum Emissions Mask—SEM)
Sets the control mode for the selected marker to Normal (see “Select Marker” on page 483). If the marker is off, a single marker is activated and placed on the screen to the right of the measurement graph. You can then adjust the trace point of the marker using the knob, numeric keypad, or keys.
See “Marker (Spectrum Emissions Mask—SEM)” on page 483 for the mode command.
Example: CALC:SEM:MARK:STAT ON selects marker 1 and sets it to Normal.
Turns off the selected marker. In addition, Off also turns off functions related to the selected marker.
See “Select Marker” on page 483 for the command to select the control mode.
Example: CALC:SEM:MARK3:STAT OFF selects marker 3 and sets it to Off.

Meas Setup (Spurious Emissions)
When the spurious emissions measurement has been selected in the Measure menu of the Spectrum Analysis mode, this key displays the appropriate measurement setup menu.
The spurious emissions measurement identifies and determines the power level of spurious emissions in certain frequency bands.
There is no equivalent remote command.
Press Avg Number to On to specify the number of measurements that will be averaged when calculating the measurement result. The average will be displayed at the end of each sweep. Off disables the measurement averaging.
State Saved: Saved in instrument state.
[:SENSe]:SPURious:AVERage:COUNt
[:SENSe]:SPURious:AVERage:COUNt?
[:SENSe]:SPURious:AVERage[:STATe] OFF|ON|0|1 turns the averaging on or off.
[:SENSe]:SPURious:AVERage[:STATe]?

Meas Setup (Spurious Emissions)
Selects the type of termination control used for the averaging function (Exp or Repeat). This determines the averaging action after the specified number of measurements (average count) is reached.
State Saved: Saved in instrument state.
[:SENSe]:SPURious:AVERage:TCONrol EXPonential|REPeat
[:SENSe]:SPURious:AVERage:TCONrol?
5.27.3 Range Table
Enables you to enter the settings for up to 20 ranges, either using the instrument front panel keys or remotely. Upon entering the range table, the measurement stops, then the analyzer is set to a constantly sweeping idle state. The analyzer is then set to the current values for range 1 (whether range 1 is on or off). If a range is currently off, the values in the range table for that range are replaced with --- (see Range 5 in the illustration) to

Meas Setup (Spurious Emissions)
indicate this range is currently inactive.

Saved State: All values for all ranges are saved in instrument state.
Sweep Time (Auto), Absolute Limit = −50 dBm, Peak Threshold = −90 dBm, Peak Excursion = 6 dB, and Abs Stop Limit Mode (Auto) is set to −50 dBm. The table below defines the other parameter’s settings for the measurement. The values for ranges 1 through 5 are based on the W-CDMA (3GPP) BTS Spurious Emission measurement.
Table 5-6 Range Table Default Settings
Range Start Freq (GHz) Stop Freq (GHz) Res BW (kHz)
1 1.920 1.980 1200
2 1.894 1.920 510
3 2.100 2.102 100
4 2.175 2.180 100
5 0.800 1.000 4000
6-20 1.5 2.5 Auto
There is no equivalent remote command.

Meas Setup (Spurious Emissions)
5.27.3.1 Range
Selects a range and updates the values on the other Range Table keys so that they reflect the settings for the selected range. If Range is set to On it is used as part of the measurement; when set to Off it is excluded.
Factory Preset: See “Range Table” on page 486.
This parameter can receive up to 20 values. The location in the list sent corresponds to the range the value is associated with. Missing values are not permitted. For example, if you want to change values 2 and 6, you must send all values up to 6. Subsequent values remain as they were.
[:SENSe]:SPURious[:RANGe][:LIST]:STATe OFF|ON|0|1
The query for this parameter will always return 20 values.
[:SENSe]:SPURious[:RANGe][:LIST]:STATe?
5.27.3.2 Start Freq
Used to set the start frequency of the analyzer.
Factory Preset: See “Range Table” on page 486.
Step Key Increment: 1, 1.5, 2, 3, 5, 7.5, 10,...
Range: Frequency range of your analyzer.
This parameter can receive up to 20 values. The location in the list sent corresponds to the range the value is associated with. Missing values are not permitted, for example, if you want to change values 2 and 6, you must send all values up to 6. Subsequent values will remain as they were.
[:SENSe]:SPURious[:RANGe][:LIST]:FREQuency:STARt
The query for this parameter will always return 20 values.

Meas Setup (Spurious Emissions)
[:SENSe]:SPURious[:RANGe][:LIST]:FREQuency:STARt?

Meas Setup (Spurious Emissions)
5.27.3.3 Stop Freq
Stop Freq is used to set the stop frequency of the analyzer.
Factory Preset: See “Range Table” on page 486.
Step Key Increment: 1, 1.5, 2, 3, 5, 7.5, 10,...
Range: Frequency range of your analyzer.
This parameter can receive up to 20 values. The location in the list sent corresponds to the range the value is associated with. Missing values are not permitted, for example, if you want to change values 2 and 6, you must send all values up to 6. Subsequent values will remain as they were.
[:SENSe]:SPURious[:RANGe][:LIST]:FREQuency:STOP
The query for this parameter will always return 20 values.
[:SENSe]:SPURious[:RANGe][:LIST]:FREQuency:STOP?
5.27.3.4 Res BW
Res BW is used to set the resolution bandwidth of the analyzer. When Auto is selected the analyzer determines the optimum setting, while Man enables you to determine the setting.
Factory Preset: See “Range Table” on page 486.
Knob Increment: Steps through the available resolution bandwidth filters.
Step Key Increment: Steps through the available resolution bandwidth filters.
Range: Resolution bandwidth range of your analyzer.
This parameter can receive up to 20 values. The location in the list sent corresponds to the range the value is associated with. Missing values are not permitted, for example, if you want to change values 2 and 6, you must send all values up to 6. Subsequent values will remain as they were.
[:SENSe]:SPURious[:RANGe][:LIST]:BANDwidth[:RESolution]:AUTO OFF|ON|0|1 selects the mode.
[:SENSe]:SPURious[:RANGe][:LIST]:BANDwidth[:RESolution]
The query for this parameter always returns 20 values.

Meas Setup (Spurious Emissions)
[:SENSe]:SPURious[:RANGe][:LIST]:BANDwidth[:RESolution]:AUTO?
[:SENSe]:SPURious[:RANGe][:LIST]:BANDwidth[:RESolution]?
5.27.3.5 Video BW
Video BW is used to set the video bandwidth of the analyzer. When Auto is selected the analyzer determines the optimum setting, while Man enables you to determine the setting.
Factory Preset: See “Range Table” on page 486.
Knob Increment: Steps through the available video bandwidth filters.
Step Key Increment: Steps through the available video bandwidth filters.
Range: Video bandwidth range of your analyzer.
This parameter can receive up to 20 values. The location in the list sent corresponds to the range the value is associated with. Missing values are not permitted, for example, if you want to change values 2 and 6, you must send all values up to 6. Subsequent values will remain as they were.
[:SENSe]:SPURious[:RANGe][:LIST]:BWIDth|BANDwidth:VIDeo:AUTO OFF|ON|0|1 selects the mode.
[:SENSe]:SPURious[:RANGe][:LIST]:BWIDth|BANDwidth:VIDeo
The query for this parameter will always return 20 values.
[:SENSe]:SPURious[:RANGe][:LIST]:BWIDth|BANDwidth:VIDeo:AUTO?
[:SENSe]:SPURious[:RANGe][:LIST]:BWIDth|BANDwidth:VIDeo?
5.27.3.6 Sweep Time
Sweep Time is used to set the sweep time of the analyzer. When Auto is selected the analyzer determines the optimum setting, while Man enables you to determine the setting.
Factory Preset: See “Range Table” on page 486.
Step Key Increment: 1, 1.5, 2, 3, 5, 7.5, 10,...
Range: Sweep time range of your analyzer.

Meas Setup (Spurious Emissions)
[:SENSe]:SPURious[:RANGe][:LIST]:SWEep:TIME:AUTO OFF|ON|0|1, OFF|ON|0|1
[:SENSe]:SPURious[:RANGe][:LIST]:SWEep:TIME:AUTO?
[:SENSe]:SPURious[:RANGe][:LIST]:SWEep:TIME: , ,... (up to 20 sweeptimes)
[:SENSe]:SPURious[:RANGe][:LIST]:SWEep:TIME?

Meas Setup (Spurious Emissions)
list sent corresponds to the range the value is associated with. Missing values are not permitted, for example, if you want to change values 2 and 6, you must send all values up to 6. Subsequent values will remain as they were.
The query for this parameter will always return 20 values.
[:SENSe]:SPURious[:RANGe][:LIST]:SWEep:TIME: 4ms,10ms
5.27.3.7 Abs Start Limit
Abs Start Limit is used to set the absolute start limit to the set value, then sets the absolute stop limit to the same value as the start limit. If any spurs are located above the current setting of Peak Excursn will be added to the results table. A red F will be appended to the amplitude value of the spur if the measured amplitude of the spur is above the limit set with Abs Start Limit.
Dependencies/ Couplings: If the limit is set to Auto, Abs Start Limit and Abs Stop Limit are coupled to make a flat limit line. If set to Man, the absolute start and stop limits can take different values to make a sloped limit line.
Factory Preset: See “Range Table” on page 486.
:CALCulate:SPURious[:RANGe][:LIST]:LIMit:ABSolute[:UPPER]:DATA:STARt
:CALCulate:SPURious[:RANGe][:LIST]:LIMit:ABSolute[:UPPER]:DATA:STARt?
Remote Command Notes: This parameter can receive up to 20 values. The location in the list sent corresponds to the range the value is associated with. Missing values are not permitted, for example, if you want to change values 2 and 6, you must send all values up to 6. Subsequent values will remain as they were.
The query for this parameter will always return 20 values.
Related commands are [:SENSe]:SPURious[:RANGe][:LIST]:SWEep:TIME:AUTO?
[:SENSe]:SPURious[:RANGe][:LIST]:SWEep:TIME?

Meas Setup (Spurious Emissions)

Meas Setup (Spurious Emissions)
5.27.3.8 Abs Stop Limit
Abs Stop Limit is used to determine the limit above which spurs will report a failure. If any spurs are located above the current setting of Peak Excursn will be added to the results table. A red F will be appended to the amplitude value of the spur if the measured amplitude of the spur is above the limit set with Abs Stop Limit.
Dependencies/ Couplings: If the limit is set to Auto, Abs Start Limit and Abs Stop Limit are coupled to make a flat limit line. If set to Man, the absolute start and stop limits can take different values to make a sloped limit line.
Factory Preset: See “Range Table” on page 486.
:CALCulate:SPURious[:RANGe][:LIST]:LIMit:ABSolute[:UPPER]:DATA:STOP ,
:CALCulate:SPURious[:RANGe][:LIST]:LIMit:ABSolute[:UPPER]:DATA:STOP?
:CALCulate:SPURious[:RANGe][:LIST]:LIMit:ABSolute[:UPPER]:DATA:STOP:AUTO OFF|ON|0|1, OFF|ON|0|1
:CALCulate:SPURious[:RANGe][:LIST]:LIMit:ABSolute[:UPPER]:DATA:STOP:AUTO?
Remote Command Notes: This parameter can receive up to 20 values. The location in the list sent corresponds to the range the value is associated with. Missing values are not permitted, for example, if you want to change values 2 and 6, you must send all values up to 6. Subsequent values will remain as they were.
The query for this command will always return 20 values.

Meas Setup (Spurious Emissions)
5.27.3.9 Peak Excursn
Peak Excursn sets the minimum amplitude variation of signals that can be identified as peaks. For example, if a value of 6 dB is selected, peaks that rise and fall more than the 6 dB above the peak threshold value are identified.
Factory Preset: See “Range Table” on page 486.
This parameter can receive up to 20 values. The location in the list sent corresponds to the range the value is associated with. Missing values are not permitted, for example, if you want to change values 2 and 6, you must send all values up to 6. Subsequent values will remain as they were.
[:SENSe]:SPURious[:RANGe][:LIST]:PEAK:EXCursion
The query for this parameter will always return 20 values.
[:SENSe]:SPURious[:RANGe][:LIST]:PEAK:EXCursion?
5.27.3.10 Pk Threshold
Pk Threshold sets the minimum amplitude of signals that can be identified as peaks. For example, if a value of -90 dBm is selected, only peaks that rise and fall more than the peak excursion value which is above -90 dBm are identified.
Factory Preset: See “Range Table” on page 486.
This parameter can receive up to 20 values. The location in the list sent corresponds to the range the value is associated with. Missing values are not permitted, for example, if you want to change values 2 and 6, you must send all values up to 6. Subsequent values will remain as they were.
[:SENSe]:SPURious[:RANGe][:LIST]:PEAK:THReshold

Meas Setup (Spurious Emissions)
The query for this parameter will always return 20 values.
[:SENSe]:SPURious[:RANGe][:LIST]:PEAK:THReshold?
Specifies the measurement type (Examine or Full). This parameter is coupled to the average mode. If you select Examine, the measurement sets Avg Mode to Exp. If you select Full, the measurement sets Avg Mode to Rep. The behavior of each measurement type is described below.
No Spurs Spurs No Spurs Spurs
Examine All active ranges are measured.
All active ranges are measured and the found spurs are reported.
All active ranges are measured.
All active ranges are measured and the found spurs are reported.
Upon completion, the measurement is set to the idle state.
Upon completion, the measurement is set to the idle state.
Upon completion, the analyzer remains set to the last range checked with an active trace.
Upon completion the analyzer is set to the range containing the worst spur found and continually sweeps this range.
The message No Spurs is displayed.
The trace containing the worst spur is restored.
The message No Spurs is displayed.
Spur is enabled (no longer greyed out).
Spur is enabled (no longer greyed out).
A marker is added, set to the frequency of the worst spur.
A marker added, set to the frequency of the worst spur.
Full All active ranges are measured.
All active ranges are measured and the found spurs are reported.
Measurement continually cycles through all active ranges.
All active ranges are measured and the found spurs are reported.
Upon completion, the measurement is set to the idle state.
Upon completion, the measurement is set to the idle state.
On each cycle of the active ranges, the report on spurs found is reset. This ensures that remote queries retrieve the trace data that matches the currently displayed results.
The message No Spurs is displayed.
The trace of the last active range is displayed.
State Saved: Saved in instrument state.
Single Continuous

Meas Setup (Spurious Emissions)
[:SENSe]:SPURious:TYPE EXAMine|FULL
[:SENSe]:SPURious:TYPE?
Remote Command Notes n/a
Enables you to view any spurs that have been found. The measurement sets the analyzer to the range in which the currently selected spur was found. The range settings changes only if the selected spur is in the range that is different from the current range settings. A marker identifies the currently selected spur on the trace.
Saved State: No values are saved to state.
[:SENSe]:SPURious:SPUR
[:SENSe]:SPURious:SPUR?
Pressing Ref Level sets the reference level of the analyzer.
Saved State: Ref Level state is saved to instrument state. Ref Level value is recalculated for each active range the value is not stored in instrument state.
Step Key

Meas Setup (Spurious Emissions)
[:SENSe]:SPURious:POWer[:RF]:LEVel
[:SENSe]:SPURious:POWer[:RF]:LEVel?
[:SENSe]:SPURious:POWer[:RF]:RANGe:AUTO OFF|ON|0|1
[:SENSe]:SPURious:POWer[:RF]:RANGe:AUTO?
5.27.7 Fast Spurious Meas
Pressing Fast Spurious Meas turns the fast spurious measurement test on or off. Pressing Fast Spurious Meas to On provides a faster method of execution as with fast spurious testing spurs above the limit line are reported. Any spurs reported outside the limit will cause the measurement to fail. See “Abs Start Limit” on page 493 for more information.
Saved State: Ref Level state is saved to instrument state. Ref Level value is recalculated for each active range the value is not stored in instrument state.
[:SENSe]:SPURious:FSMeas OFF|ON|0|1
[:SENSe]:SPURious:FSMeas?

Meas Setup (Spurious Emissions)

Programming Fundamentals

Programming Fundamentals · Programming Fundamentals

This section is not intended to teach you everything about the SCPI (Standard Commands for Programmable Instruments) programming language. The SCPI Consortium or IEEE can provide that level of detailed information.
Topics covered in this chapter include:
For more information refer to:
IEEE Standard 488.1-1987, IEEE Standard Digital Interface for Programmable Instrumentation. New York, NY, 1998.
IEEE Standard 488.2-1987, IEEE Standard Codes, Formats, Protocols and Comment Commands for Use with ANSI/IEEE Std488.1-1987. New York, NY, 1998.
A typical command is made up of keywords set off by colons. The keywords are followed by parameters that can be followed by optional units.
The instrument does not distinguish between upper and lower case letters. In the documentation, upper case letters indicate the short form of the keyword. The lower case letters, indicate the long form of the keyword. Either form may be used in the command.
is the same as SENSE:FREQ:start 1.5 MHz
Programming Fundamentals · SCPI Language Basics · Programming Fundamentals

Commands are not case sensitive and there are often many different ways of writing a particular command. These are examples of valid commands for a given command syntax:
Command Syntax Sample Valid Commands
[SENSe:]BANDwidth[:RESolution] The following sample commands are all identical. They will all cause the same result.
MEASure:SPECtrum[n]? • MEAS:SPEC?
The number 3 in the last meas example causes it to return different results then the commands above it. See the command description for more information.
[:SENSe]:DETector[:FUNCtion] NEGative|POSitive|SAMPle
INITiate:CONTinuous ON|OFF|1|0 The sample commands below are identical.
Programming Fundamentals · SCPI Language Basics · Programming Fundamentals

Special Character Meaning Example
| A vertical stroke between parameters indicates alternative choices. The effect of the command is different depending on which parameter is selected.
The choices are external, internal, and line. Ex: TRIG:SOURCE INT
is one possible command choice.
A vertical stroke between keywords indicates identical effects exist for both keywords. The command functions the same for either keyword. Only one of these keywords is used at a time.
[ ] keywords in square brackets are optional when composing the command. These implied keywords will be executed even if they are omitted.
The following commands are all valid and have identical effects: Ex1: bandwidth:auto Ex2: band:resolution:auto Ex3: sense:bandwidth:auto
< > Angle brackets around a word, or words, indicates they are not to be used literally in the command. They represent the needed item.
In this command example the word should be replaced by an actual frequency.
Ex: SENS:FREQ 9.7MHz.
{ } Parameters in braces can optionally be used in the command either not at all, once, or several times.
Programming Fundamentals · SCPI Language Basics · Programming Fundamentals

There are four basic types of parameters: booleans, keywords, variables and arbitrary block program data.
OFF|ON|0|1 (Boolean) This is a two state boolean-type parameter. The numeric value 0 is equivalent to OFF. Any numeric value other than 0 is equivalent to ON. The numeric values of 0 or 1 are commonly used in the command instead of OFF or ON. Queries of the parameter always return a numeric value of 0 or 1.
keyword The keywords that are allowed for a particular command are defined in the command syntax description.
Units Numeric variables may include units. The valid units for a command depend on the variable type being used. See the following variable descriptions. The indicated default units will be used if no units are sent. Units can follow the numerical value with, or without, a space.
Variable A variable can be entered in exponential format as well as standard numeric format. The appropriate range of the variable and its optional units are defined in the command description.
The following keywords may also be used in commands, but not all commands allow keyword variables.
The numeric value for the function’s MINimum, MAXimum, or DEFault can be queried by adding the keyword to the command in its query form. The keyword must be entered following the question mark.
Programming Fundamentals · SCPI Language Basics · Programming Fundamentals

Is a positive rational number followed by optional units. The default unit is Hz. Acceptable units include: HZ, KHZ, MHZ, GHZ.
Is a rational number followed by optional units. The default units are seconds. Acceptable units include: S, MS, US.
Is a rational number followed by optional units. The default units are V. Acceptable units include: Volts, V, MV, UV.
Is a rational number followed by optional units. The default units are dBm. Acceptable units include: DBM, DBMV, W.
Is a positive rational number followed by optional units. The default units are dB. Acceptable units include: DB.
Is a rational number followed by optional units. The default units are degrees. Acceptable units include: DEG, RAD.
An integer value has no units.
Is a floating point number, with no units.
Is a rational number between 0 and 100, with no units.
Is a series of alpha numeric characters.
Specifies a series of bits rather than a numeric value. The bit series is the binary representation of a numeric value. There are no units.
Bit patterns are most often specified as hexadecimal numbers, though octal, binary or decimal numbers may also be used. In the SCPI language these numbers are specified as:

Some parameters consist of a block of data. Block data There are a few standard types of block data. Arbitrary blocks of program data can also be used.
Is an array of rational numbers corresponding to displayed trace data. See FORMat:DATA for information about available data formats.
A SCPI command often refers to a block of current trace data with a variable name such as: Trace1, TRACE2, or trace3, depending on which trace is being accessed.
Consists of a block of data bytes. The first
information sent in the block is an ASCII header beginning with #. The block is terminated with a semi-colon. The header can be used to determine how many bytes are in the data block. There are no units.
For example, suppose the header is #512320.
Multiple commands can be written on the same line, reducing your code space requirement. To do this:

SCPI Termination and Separator Syntax
A terminator must be provided when an instrument is controlled using RS-232. There are several issues to be understood about choosing the proper SCPI terminator and separator when this is the case. There is no current SCPI standard for RS-232. Although one intent of SCPI is to be interface independent, is only defined for IEEE 488 operation. At the time of this writing, the RS-232 terminator issue was in the process of being addressed in IEEE standard 1174.
A semicolon (;) is not a SCPI terminator, it is a separator. The purpose of the separator is to queue multiple commands or queries in order to obtain multiple actions and/or responses. Make sure that you do not attempt to use the semicolon as a terminator when using RS-232 control.
All binary trace and response data is terminated with , as defined in Section 8.5 of IEEE Standard 488.2-1992, IEEE Standard Codes, Formats, Protocols and Common Commands for Use with ANSI/IEEE Std 488.1-1987. New York, NY, 1992.
The following are some examples of good and bad commands. The examples are created from a theoretical instrument with the simple set of commands indicated below:
[:SENSe]
:POWer
[:RF] :ATTenuation 40dB
:TRIGger
[:SEQuence] :EXTernal [1]
:SLOPe

[:SENSe]
:FREQuency
:STARt :POWer [:RF]
:MIXer
:RANGe [:UPPer]
Bad Command Good Command
PWR:ATT 40dB POW:ATT 40dB
The short form of POWER is POW, not PWR.
FREQ:STAR 30MHz;MIX:RANG -20dBm
FREQ:STAR 30MHz;POW:MIX:RANG -20dBm
The MIX:RANG command is in the same :SENSE subsystem as FREQ, but executing the FREQ command puts you back at the SENSE level. You must specify POW to get to the MIX:RANG command.
FREQ:STAR 30MHz;POW:MIX RANG -20dBm
FREQ:STAR 30MHz;POW:MIX:RANG -20dBm
MIX and RANG require a colon to separate them.
:POW:ATT 40dB;TRIG:FREQ:STAR 2.3GHz :POW:ATT 40dB;:FREQ:STAR 2.3GHz
:FREQ:STAR is in the :SENSE subsystem, not the :TRIGGER subsystem.
:POW and FREQ are within the same :SENSE subsystem, but they are two separate commands, so they should be separated with a semicolon, not a colon.
:POW:ATT -5dB;:FREQ:STAR 10MHz :POW:ATT 5dB;:FREQ:STAR 10MHz
Attenuation cannot be a negative value.
Programming Fundamentals · SCPI Language Basics · Programming Fundamentals

There are a number of things you can do in your programs to make them run faster:
“Turn off the display updates.” on page 511
“Use binary data format instead of ASCII” on page 512
“Minimize the number of GPIB transactions.” on page 512
“Put ADC Ranging in Bypass for FFT Measurements” on page 513
“Minimize DUT/instrument setup changes.” on page 513
“Consider using LAN instead of GPIB.” on page 514
There are additional things you can do to run faster if you are using a measurement personality option (i.e. instrument Modes other than the standard Spectrum Analysis Mode). These considerations only apply to specific option modes.
Turn off the display updates.
:DISPlay:ENABle OFF turns off the display. That is, the data may still be visible, but it will no longer be updated. Updating the display slows down the measurement. For remote testing, since the computer is processing the data rather than a person, there is no need to display the data on the analyzer screen.
Programming Fundamentals · Improving Measurement Speed · Programming Fundamentals

The ASCII data format is the instrument default since it is easier for people to understand and is required by SCPI for *RST. However, data input/output is faster using the binary formats.
:FORMat:DATA REAL,64 selects the 64-bit binary data format for all your numerical data queries. You may need to swap the byte order if you are using a PC rather than UNIX. NORMal is the default byte order. Use :FORMat:BORDer SWAP to change the byte order so that the least significant byte is sent first. (Real,32 which is smaller and somewhat faster, should only be used if you don’t need full resolution for your data. Some frequency data may require full 64 bit resolution.)
When using the binary format, data is sent in a block of bytes with an
ASCII header. A data query would return the block of data in the following format: #DNNN
To parse the data:
For example, suppose the header is #512320.
Minimize the number of GPIB transactions.
When you are using the GPIB for control of your instrument, each transaction requires driver overhead and bus handshaking, so minimizing these transactions reduces the time used.
You can reduce bus transactions by sending multiple commands per transaction. See the information on “Putting Multiple Commands on the Same Line” in the SCPI Language Basics section.
If you are using the MEASURE key measurements and are making the same measurement multiple times with small changes in the measurement setup, use the single READ command. It is faster then using INITiate and FETCh.
Programming Fundamentals · Improving Measurement Speed · Programming Fundamentals

Avoid unnecessary use of *RST.
Remember that while *RST does not change the current Mode, it presets all the measurements and settings to their factory defaults. This forces you to reset your analyzer’s measurement settings even if they use similar mode setup or measurement settings. See Minimize DUT/instrument setup changes. below. (Also note that *RST may put the instrument in single measurement/sweep for some modes.)
Setting ADC ranging to the Bypass mode can speed FFT measurements up by 10% to 50%. (Use ADC:RANG NONE) Bypass allows triggered FFT measurements to occur at the trigger time instead of following an autoranging time, so it can improve measurement speed. It does, however, add additional noise degrading your signal to noise level, so it should be used carefully.
Minimize DUT/instrument setup changes.
Programming Fundamentals · Improving Measurement Speed · Programming Fundamentals

Consider using LAN instead of GPIB.
LAN allows faster I/O of data, especially if you are moving large blocks of data. You will not get this improved throughput if there is excessive LAN traffic (i.e. your test instrument is connected to enterprise LAN). You may want to use a private LAN that is only for your test system.
Using an Option Mode: Minimize the number of GPIB transactions.
When you are using the GPIB for control of your instrument, each transaction requires driver overhead and bus handshaking, so minimizing these transactions reduces the time used.
The CONFigure/MEASure/READ commands for measurements in the option Modes allow you to send center frequency setup information along with the command. (for example, MEAS:PVT? 935.2MHz) This sets the power vs. time measurement to it’s defaults, then changes the center frequency to 935.2 MHz, initiates a measurement, waits until it is complete and returns the measurement data.
For example, send READ:PFER? ;PFER? ;PFER? This single transaction initiates three different phase and frequency error measurements at each of the three different frequencies provided and returns the data. Then you read the three sets of data.
Programming Fundamentals · Improving Measurement Speed · Programming Fundamentals

Using an Option Mode: Avoid automatic attenuator setting.
The internal process for automatically setting the value of the attenuator requires measuring an initial burst to identify the proper attenuator setting before the next burst can be measured properly. If you know the amount of attenuation or the signal level needed for your measurement, just set it.
Using an Option Mode: Optimize your GSM output RF spectrum switching measurement.
For ORFS (switching), setting the break frequency to zero (0) puts the analyzer in a measurement setup where it can use a direct time measurement algorithm, instead of an FFT-based algorithm. This non-FFT approach is faster. (However, remember that your break frequency for ORFS (modulation) measurements must be >400 kHz for valid measurements, so you will need to change the break frequency if you are making both types of measurements.)
Using an Option Mode: Avoid using RFBurst trigger for single burst signals.
RFBurst triggering works best when measuring signals with repetitive bursts. For a non-repetitive or single burst signals, use the IF(video) trigger or external trigger, depending on what you have available.
RFBurst triggering depends on its establishment of a valid triggering reference level, based on previous bursts. If you only have a single burst, the peak detection nature of this triggering function, may result in the trigger being done at the wrong level/point generating incorrect data, or it may not trigger at all.
Are you making a single burst measurement?
To get consistent triggering and good data for this type of measurement application, you need to synchronize the triggering of the DUT with the analyzer. You should use the analyzer’s internal status system for this.
The first step in this process is to initialize the status register mask to look for the “waiting for trigger” condition (bit 5). Use :STATus:OPERation:ENABle 32
Then, in the measurement loop:
Programming Fundamentals · Improving Measurement Speed · Programming Fundamentals

Make sure the attenuation is set manually. Do NOT use automatic attenuation as this requires an additional burst to determine the proper attenuation level before the measurement can be made.
This process cannot be done by using with the current VXI plug-n-play driver implementation. You will need to use the above SCPI commands.
Using an Option Mode: When making power measurements on multiple bursts or slots, use CALCulate:DATA<n>:COMPress?
The CALC:DATA:COMP? query is the fastest way to measure power data for multiple bursts/slots. There are two reasons for this: 1. it can be used to measure data across multiple, consecutive slots/frames with just one measurement, instead of a separate measurement on each slot, and 2. it can pre-process and/or decimate the data so that you only return the information that you need which minimizes data transfer to the computer.
For example: let’s say you want to do a power measurement for a GSM base station where you generate a repeating frame with 8 different power levels. You can gather all the data with a single CALC:DATA:COMP? acquisition, using the waveform measurement.
With CALC:DATA2:COMP? MEAN,9,197,1730 you can measure the mean power in those bursts. This single command will measure the data across all 8 frames, locate the first slot/burst in each of the frames, calculate the mean power of those bursts, then return the resulting 8 values.
Let’s set up the GSM Waveform measurement:
Programming Fundamentals · Improving Measurement Speed · Programming Fundamentals

There are two versions of this command depending on your firmware revision. Earlier revisions require the optional variables be entered in terms of their position in the trace data array. Current instruments allow the variables to be entered in terms of time.
For early firmware revisions you need to know the sample interval. In the waveform measurement it is equal to the aperture value. Query :WAVeform:APERture? to find the sample interval. (Note: the WAV:APER? command always takes decimation into account.) The sample interval (aperture value) is dependent on the settings for resolution bandwidth, filter type, and decimation. See the following table to see how these value relate.
The parameters for this GSM example are: MEAN,9,197,1730 (or with later firmware: MEAN,25us,526us,579.6us,8)
Programming Fundamentals · Improving Measurement Speed · Programming Fundamentals

Table 6-1 GSM Parameters for 1 Slot/Frame Measurement Requirements
Decimation Aperture Start Length Repeat
500 or 300 kHz
4 or 1 dependent on settings
24 µseca 526 µseca 576.9 µseca
500 kHz Gaussian 1 0.2 µsec 124 2630 2884.6
500 kHz Gaussian 4 0.8 µsec 31 657 721.15
500 kHz Flat 1 0.4 µsec 61 1315 1442.3
500 kHz Flat 4 1.6 µsec 15 329 360.575
300 kHz Gaussian 1 0.2667 µsec 90 1972 2163.1
300 kHz Gaussian 4 1.07 µsec 22 492 539.16
300 kHz Flat 1 0.6667 µsec 36 789 865.31
300 kHz Flat 4 2.667 µsec 9 197 216.33
a. The use of time values is only allowed in firmware versions of A.02.00 and later.

Programming Command Compatibility Across Model Numbers and Across Modes
When you select different modes you get different sets of available programming commands. That is, only the commands that are appropriate for the current mode are available. Also, some commands have the same syntax in different modes but have different ranges or settings that are only appropriate to the current mode.
The following table shows which command subsystems are the same across different modes. If there is no “X” by a particular subsystem, then the set of available commands is different in those modes. Command ranges or defaults may also be different. Refer to the programming command descriptions in the documentation for each mode for details.
Command Subsystem Same command set is available:
Same command set is available:
SA mode compared with the application modes: W-CDMA, cdmaOne, cdma2000, 1xEV-DO, Basic, GSM, EDGE, NADC, or PDC
SA mode compared with the application mode: Phase Noise
IEEE common commands X X
COUPle not available in these application modes
not available in this application modes
INPut not available in these application modes

Programming Command Compatibility Across Model Numbers and Across Modes
Command Subsystem Same command set is available:
Same command set is available:
SA mode compared with the application modes: W-CDMA, cdmaOne, cdma2000, 1xEV-DO, Basic, GSM, EDGE, NADC, or PDC
SA mode compared with the application mode: Phase Noise
MMEMory:STORe:TRACe not available in application modes
[SENSe:]CHANnel
[SENSe:]CORRection
[SENSe:]FEED
[SENSe:]FREQuency:CENTer X
[SENSe:]FREQuency:
not available in application modes
not available in application modes
[SENSe:]
[SENSe:]POWer
[SENSe:]RADio
[SENSe:]SYNC
TRACe not available in application modes

Programming Command Compatibility Across Model Numbers and Across Modes
Some programming commands operate differently depending on which Mode the analyzer is set to.
Basic, cdmaOne, cdma2000, 1xEV-DO, W-CDMA, GSM, EDGE, NADC, PDC Modes
Spectrum Analysis, Phase Noise and Noise
Accesses the measurement and sets the instrument settings to the defaults. Averaging is turned on and set to 10. The instrument is put in single measurement mode. It does not initiate a measurement. Use INIT:IMM to make one measurement.
Accesses the measurement and sets the instrument settings to the defaults. If you were already in single measurement mode, it takes one measurement and then waits. If you were in continuous measurement mode it continues to measure.
*ESE default Default is 255 which means that every error/status bit change that has occurred will be returned with a *ESR? query. You must set the value of *ESE to choose only the bits/status that you want returned.
Default is 0 which means that none of the error/status bit changes that have occurred will be returned with a *ESR? query. You must set the value of *ESE to choose the bits/status that you want returned.
For these modes, only one trigger source can be selected and it will be common across the modes. Also, only one value can be set for the trigger delay, level, or polarity.
For these modes, a unique trigger source can be selected for each mode. Also, each trigger source can have unique settings for the its delay, level, and polarity.
Saving and recalling traces
Traces can only be saved when in the Spectrum Analysis mode (MMEM:STOR:TRAC). This is because the instrument state must be saved along with the trace data and the state data varies depending on the number of modes currently available in the instrument.

Programming Command Compatibility Across Model Numbers and Across Modes
*RST Resets instrument, putting it in continuous measurement mode. Use INIT:CONT OFF to select single measurement mode and INIT:IMM to start one measurement.
Resets instrument, putting it in single measurement mode. One measurement is initiated when the command is sent.
Resets instrument, putting it in single measurement mode. No measurement is initiated when the command is sent. Use INIT:IMM to start one measurement.
Accesses the measurement and sets the instrument settings to the defaults. If you were already in single measurement mode, it takes one measurement and then waits.
Same as PSA.
Accesses the measurement and sets the instrument settings to the defaults. If you were already in single measurement mode, it does not initiate a measurement. Use INIT:IMM to make one measurement.
Accesses the measurement and sets the instrument settings to the defaults. If you were already in single measurement mode, it takes one measurement and then waits.
*ESE default Default is 255 which means that every error/status bit change that has occurred will be returned with a *ESR? query. You must set the value of *ESE to choose only the bits/status that you want returned.
Default is 0 which means that none of the error/status bit changes that have occurred will be returned with a *ESR? query. You must set the value of *ESE to choose the bits/status that you want returned.
Same as VSA A.04.00.
Default is 0 which means that none of the error/status bit changes that have occurred will be returned with a *ESR? query. You must set the value of *ESE to choose the bits/status that you want returned.
*LRN The command is not available.
The command is available.
The command is available.
In Spectrum Analysis mode only one value can be set for the trigger’s source, delay, level, or polarity.
You can select a unique trigger source for each mode. Each trigger source can have unique settings for the its delay, level, and polarity.
Same as VSA A.04.00.
You can select a unique trigger source for each mode. Each trigger source can have unique settings for the its delay, level, and polarity.
Basic, GSM, EDGE, cdmaOne, cdma2000, W-CDMA, NADC, PDC modes function the same as VSA
Programming Fundamentals · TRIGger commands · Programming Fundamentals

Programming Command Compatibility Across Model Numbers and Across Modes
AUTO ON|OFF control and setting manual values
We recommend that you set a function’s automatic state to OFF, before you send it your manual value.
We recommend that you set a function’s automatic state to OFF, before you send it your manual value.
We recommend that you set a function’s automatic state to OFF, before you send it your manual value.
Some functions will turn off the automatic mode when you send a specific manual value, but others will not. This also varies with the instrument model.
Some functions will turn off the automatic mode when you send a specific manual value, but others will not. This also varies with the instrument model.
Some functions will turn off the automatic mode when you send a specific manual value, but others will not. This also varies with the instrument model.
Programming Fundamentals · Programming Fundamentals

Using the LAN to Control the Instrument
Refer to the function description chapters for information about configuring the instrument input/output settings from the front panel. Use the SYSTem commands to change settings remotely.
You can use the instrument LAN connection to transfer files. For example, you can use the ftp functionality to download instrument screen dumps to an external server.
The following is an example of an ftp session from an MSDOS window on a PC:
You are now in the instrument /users directory and can get files from the instrument. The ftp commands in the following steps may not all be available from your controller. To show the ftp commands available on your system, type help at the prompt. To end the ftp session, type quit.
Programming Fundamentals · Programming Fundamentals

Using the LAN to Control the Instrument
instrument operation, and for the operation of optional personality modes.
This “gets” (copies) your file. The file is copied to the location you were pointing to when you started the ftp process. To query the current location, enter lcd .(include the period). To change the current location, enter the desired path/directory location as follows: lcd C:\my path\mydir
The Standard UNIX FTP Command:
Synopsis ftp [-g] [-i] [-n] [-v] [server-host] [-B DataSocketBufferSize]
Description The ftp command is used to transfer files using the File Transfer Protocol. ftp transfers files over a network connection between a local machine and the remote server-host.
Options and Parameters When ftp is invoked with a server-host specified, a connection is opened immediately. Otherwise, ftp waits for user commands.
The following options are supported:
-g disables expansion of shell metacharacters in file and directory names
-i disables prompts during multiple-file operations
-n disables automatic log-in
-v enables verbose output
-B specifies a new DataSocketBufferSize
server-host the name or address of the remote host.
Programming Fundamentals · Programming Fundamentals

Using the LAN to Control the Instrument
This table lists the available user commands.
Table 6-2 ftp Commands
ascii Sets the file transfer type to ASCII.
binary Sets the file transfer type to binary.
bye Closes the connection to the host and exits ftp.
cd remote_directory Sets the working directory on the host to remote_directory.
delete remote_file Deletes remote_file or empty remote_directory.
dir [remote_directory]
Lists the contents of the specified remote_directory. If remote_directory is unspecified, the contents of the current remote directory are listed.
get remote_file [local_file]
Copies remote_file to local_file. If local_file is unspecified, ftp uses the remote_file name as the local_file name.
help Provides a list of ftp commands.
help command Provides a brief description of command.
image Sets the file transfer type to binary.
lcd [local_directory] Sets the local working directory to local_directory.
ls [remote_directory]
Lists the contents of the specified remote_directory. If the remote_directory is unspecified, the contents of the current remote directory are listed.
Copy remote_file to the local system. If local_file is unspecified, ftp uses the remote_file name as the local_file name.
Copies local_file to remote file. If remote_file is unspecified, ftp uses the local_file name as the remote_file name.
put local_file [remote_file]
Copies local_file to remote file. If remote_file is unspecified, ftp uses the local_file name as the remote_file name.
quit Closes the connection to the host and exits ftp.

Using the LAN to Control the Instrument
Using telnet to send commands to your instrument works in a similar way to communicating over GPIB. You establish a connection with the instrument, and then send or receive information using SCPI commands.
On unix or PC:
The syntax of the telnet command is:
telnet
The initial telnet connection message will be displayed and then a SCPI> prompt. At the SCPI prompt, simply enter the desired SCPI commands.
On a PC (with telnet gui that has host/port setting menu):
You would type at the dos prompt
Unix Telnet Example:
To connect to the instrument with host name aaa and port number 5023, enter the following command:
When you connect to the instrument, it will display a welcome message and a command prompt. The instrument is now ready to accept your SCPI commands. As you type SCPI commands, query results appear on the next line. At any time, you can send a by pressing cntrl-c on your keyboard. When you are done, break the telnet connection using your escape character, and type quit.
When the instrument responds with the welcome message and the SCPI prompt, you can immediately enter programming (SCPI) commands. Typical commands might be:
The small program above sets the instrument to measure a signal amplitude by placing a marker on the maximum point of the trace, and then querying the instrument for the amplitude of the marker.
You need to press Enter after typing in each command. After pressing

Using the LAN to Control the Instrument
Enter on the last line in the example above, the instrument returns the amplitude level of the marker to your computer and displays it on the next line. For example, after typing CALC:MARK:MAX? and pressing Enter, the computer could display:
When you are done, close the telnet connection. Enter the escape character to get the telnet prompt. The escape character (Ctrl and "]" in this example) does not print.
At the telnet prompt, type quit or close.
The telnet connection closes and you see your regular prompt.
Connection closed.
The following example shows a terminal screen using the example commands above.
Telnet Example:
Agilent Technologies,E4440A,US41220095,A.02.04 20010921 10:52:07
SCPI>calc:mark:mode pos
SCPI>calc:mark:max
SCPI>calc:mark:x?
SCPI>
To remedy this, you need to change your telnet connection to “character-by-character” mode. This can be accomplished in most systems by escaping out of telnet to the telnet> prompt and then typing mode char. If this does not work, consult your telnet program's documentation for how to change to “character-by-character” mode.
Programming Fundamentals · Welcome to at42 · Programming Fundamentals

Using the LAN to Control the Instrument
The Standard UNIX TELNET Command:
Synopsis telnet [host [port]]
Description The telnet command is used to communicate with another host using the TELNET protocol. When telnet is invoked with host or port arguments, a connection is opened to host, and input is sent from the user to host.
Options and Parameters telnet operates in line-by-line mode or in character-at-a-time mode. In line-by-line mode, typed text is first echoed on the screen. When the line is completed by pressing the Enter key, the text line is then sent to host. In character-at-a-time mode, text is echoed to the screen and sent to host as it is typed.
In some cases, if your telnet connection is in “line-by-line” mode, there is no local echo. This means you will not be able to see the characters you are typing on your computer's display until after you press the Enter key.
To remedy this, you need to change your telnet connection to “character-by-character” mode. This can be accomplished in most systems by escaping out of telnet to the telnet> prompt and then typing mode char. Consult your telnet program's documentation for how to change to “character-by-character” mode.
Programming Fundamentals · Programming Fundamentals

Using the LAN to Control the Instrument
Your instrument implements a sockets Applications Programming Interface (API) compatible with Berkeley sockets, Winsock, and other standard sockets APIs. You can write programs using sockets to control your instrument by sending SCPI commands to a socket connection you create in your program. Refer to Using a Java™Applet Over Socket LAN in this chapter for example programs using sockets to control the instrument.
Setting Up Your Instrument for Socket Programming
Before you can use socket programming, you must identify your instrument’s socket port number. The default is 5025.
Troubleshooting help:
You can verify that you can open a socket connection to your instrument by using telnet:
telnet 5025
Characters typed from your keyboard won’t be echoed from the instrument and the SCPI prompt won’t be given. However, you will be able to send commands and query the instrument. For example, you can type *idn? and the instrument identification string will be returned.
Programming Fundamentals · Programming Fundamentals

Using the LAN to Control the Instrument
SICL LAN is a LAN protocol using the Standard Instrument Control Library (SICL). It provides control of your instrument over the LAN, using a variety of computing platforms, I/O interfaces, and operating systems. With SICL LAN, you control your remote instrument over the LAN with the same methods you use for a local instrument connected directly to the controller with the GPIB. More information about SICL LAN can be found in the HP Standard Instrument Control Library user’s guide for HP-UX, part number E2091-90004.
Your instrument implements a SICL LAN server. To control the instrument, you need a SICL LAN client application running on a computer or workstation that is connected to the instrument over a LAN. Typical applications implementing a SICL LAN client include
Older versions of National Instruments’ VISA does not support the VXI-11 Instrument Protocol. Contact National Instruments for their latest version.
SICL LAN can be used with Windows 95, Windows 98, Windows NT, and HP-UX.
Your instrument has a SICL LAN server to emulate GPIB over LAN, but it cannot be used to control other externally connected GPIB instruments.
Collecting SICL LAN Set-up Information
Before you set up your instrument as a SICL LAN server, you need some information about your instrument. The “value” of the following parameters is used to set up your VISA/SICL LAN client application:
Emulated GPIB Name The GPIB name is the name given to a device used to communicate with the instrument. Your instrument is shipped with gpib7 as its GPIB name. The GPIB name is the same as the remote SICL address.
Emulated GPIB Logical Unit The logical unit number is a unique integer assigned to the device to be controlled using SICL LAN. Your instrument is shipped with the logical unit number set
Programming Fundamentals · Programming Fundamentals

Using the LAN to Control the Instrument
to 8.
This can’t be change, but you don’t care. Numbers 0 through 30, excluding 21, are valid logical unit numbers for your instrument. Logical unit number 21 is used for the instrument’s internal emulation mode. (If you are using Agilent VEE and SICL LAN, the logical unit number is limited to the range of 0-8.)
Emulated GPIB Address The emulated GPIB address (bus address) is assigned to the device to be controlled using SICL LAN. The instrument is shipped with the emulated GPIB address set to 18. The emulated GPIB address will always be 18, regardless of what you set the GPIB address to.
The SICL LAN server uses the GPIB name, GPIB logical unit number, and GPIB address configuration on the SICL LAN client to communicate with the client. You must match these parameters exactly (including case) when you set up the SICL LAN client and server.
Configuring Your Instrument as a SICL LAN Server
After you have collected the required information from the SICL LAN client, perform the following steps to set up your instrument as a SICL LAN server:
Press System, Config I/O, SICL Server, Emulated GPIB Name, and notice that it is gpib7.
Configuring a PC as a SICL LAN Client
The descriptions here are based on Agilent’s VISA revision G.02.02, model number 2094G. A copy of Agilent VISA instrument io libraries can be found on Agilent’s website:
http://www.agilent.com/find/iolib
http://www.agilent.com/find/vee
The VISA User’s Guide information on LAN programming may also be useful, see:
ftp://ftp.agilent.com/pub/mpusup/pc/binfiles/iop/index.html
The following assumes a LAN connection between your computer and your instrument. This will not work for the GPIB to LAN gateway.
Programming Fundamentals · see also · Programming Fundamentals

Using the LAN to Control the Instrument
10.Enter a VISA interface name, such as GPIB1.
11.Enter the host name or IP address of your instrument in the host name field, such as aaa.companyname.com or 137.12.255.755.
12.Enter a Remote SICL address, such as GPIB7.
13.Set the LAN interface to match the defined LAN client.
14.Select OK.
15.Close I/O Configuration by selecting OK.
Controlling Your Instrument with SICL LAN and HP/Agilent VEE
Before you can use SICL LAN with VEE, you need to set up VISA/SICL LAN I/O drivers for use with your VEE application. Consult your VEE documentation for information how to do this.
The logical unit number is the same as the interface select code (ISC). VEE reserves ISC values 9-18, and does not allow you to use them for SICL/LAN communications with your instrument. VEE also does not allow any ISC values higher than 18.
After you have the VISA/SICL LAN I/O drivers installed, perform the steps below to set up VEE to control your instrument:
Programming Fundamentals · Programming Fundamentals

Using the LAN to Control the Instrument
Figure 6-1 Adding Your Instrument as a VEE Device

To send SCPI commands to the instrument, select I/O|Instrument Manager, and the GPIB device just added. Select Direct I/O. You can now type SCPI commands in the command window, and they are sent over the LAN to your instrument.
Controlling Your Instrument with SICL LAN and Agilent BASIC for Windows
Before you can use Agilent BASIC for Windows with SICL LAN, you need to set up VISA/SICL LAN I/O drivers for use with your BASIC applications. Consult your BASIC documentation for information how to do this.
To set up SICL LAN for BASIC, add the following statement to your AUTOST program (all on a single line):
LOAD BIN "GPIBS;DEV lan[analyzer IP address]:GPIB name TIME 30 ISC 7"
Replace analyzer IP address with the IP address of your instrument, GPIP name with the GPIB name given to your instrument, and 7 with the logical unit number.
For example, the following LOAD statement should be added to your AUTOST program for the parameters listed below:
instrument IP address 191.108.344.225
instrument GPIB name inst0
Programming Fundamentals · Programming Fundamentals

Using the LAN to Control the Instrument
logical unit number 7
timeout value (seconds) 30
LOAD statement (all on a single line)
LOAD BIN “GPIBS;DEV lan[191.108.344.225]:inst0 TIME 30 ISC 7”
Consult your BASIC documentation to learn how to load the SICL driver for BASIC.
After the SICL driver is loaded, you control your instrument using commands such as the following:
where 18 is the device address for the instrument.
See the BASIC example program in this chapter for more information.
Controlling Your Instrument with SICL LAN and BASIC for UNIX (Rocky Mountain BASIC)
Before you can use Rocky Mountain Basic (HPRMB) with SICL LAN, you will need to set up the SICL LAN I/O drivers for HPRMB. Consult your system administrator for details.
Create a .rmbrc file in your root directory of your UNIX workstation with the following entries:
SELECTIVE_OPEN=ON Interface 8= "lan[analyzer IP address]:GPIB name";NORMAL
Replace analyzer IP address with the IP address of your instrument, and GPIB name with the GPIB name given to your instrument. Also replace the “8” of Interface 8 with the logical unit number. Consult your HPRMB documentation for the exact syntax.
After your SICL driver is configured correctly on your UNIX workstation, you control your instrument using commands such as the following:
where 18 is the device address for the instrument.
Programming Fundamentals · Programming Fundamentals

Using the LAN to Control the Instrument
(There is a VEE example program provided on the documentation CD-ROM.)
(There is a LabView example program provided on the documentation CD-ROM.)
To control your instrument via socket LAN using VEE, click on the VEE menu titled "I/O." Then select “To/From Socket” and position the I/O object box on the screen. Fill in the following fields:
For faster troubleshooting, you may want to set the timeout to a smaller number. If the host name you enter doesn't work, try using the IP address of your instrument (example: 191.108.43.5). Using the IP address rather than the hostname may also be faster. See Figure 5 for an example of an VEE screen.
Programming Fundamentals · Programming Fundamentals

Using the LAN to Control the Instrument
Figure 5 Sample VEE Screen

Programming Fundamentals · Programming Fundamentals

Using the LAN to Control the Instrument
There is a programming example in the PSA Measurement Guide and Programming Examples that demonstrates simple socket programming with Java. It is written in Java programming language, and will compile with Java compilers versions 1.0 and above.
This program is also on the documentation CD ROM that shipped with your product.
The PSA Measurement Guide and Programming Examples book contains two examples of simple LAN socket programs. They are written in C. One compiles in the HP-UX UNIX environment and one is written for the WIN32 environment.
In UNIX, LAN communication via sockets is very similar to reading or writing a file. The only difference is the openSocket() routine, which uses a few network library routines to create the TCP/IP network connection. Once this connection is created, the standard fread() and fwrite() routines are used for network communication.
In Windows, the routines send() and recv() must be used, because fread() and fwrite() may not work on sockets.
Troubleshooting the Initial Connection
Getting the instrument to work with your network often requires detailed knowledge of your local network software. This section attempts to help you with some common problems. Contact your network administrator for additional assistance.
The instrument LAN interface does not need or include any proprietary driver software. It was designed to operate with common network utilities and drivers.
Programming Fundamentals · Programming Fundamentals

Using the LAN to Control the Instrument
Either a hardware problem or a software problem can prevent the instrument's remote file server from communicating over the LAN. The following common problems may be encountered:
Communications Not Established If you have just installed and configured the LAN interface and you have never been able to access the instrument via ftp or telnet, go directly to “Pinging the Instrument from a Computer or Workstation” on page 6-542.
If you have previously been able to access the instrument via ftp or telnet and now cannot do so, check the following:
o Has any hardware been added or moved on your network? This includes adding or removing any workstations or peripherals, or changing any cabling.
o Have software applications been added to the network?
o Has the functionality been turned off from the front panel? Press System, Config I/O, SCPI LAN.
o Have any configuration files been modified? Pressing System, Restore Sys Defaults restores the original factory defaults and you will have to re-set the instrument IP address and host name.
o Is the upper- and lower-case character usage in your host name consistent?
o Have any of the following files been deleted or overwritten?
— /etc/inetd.conf
PCs:
— dependent network files
If you know or suspect that something has changed on your network, consult with your network administrator.
Timeout Errors Timeout errors such as “Device Timeout,” “File Timeout,” and “Operation Timeout,” are symptoms of one or both of the following problems:
— The currently configured timeout limits are too short compared to
the time it takes the LAN to complete some operations. This problem may occur during periods of increased LAN traffic.
— The LAN connection has failed, or fails occasionally.
To increase your timeout period, refer to your computer documentation for instructions. Contact your LAN administrator if

Using the LAN to Control the Instrument
problems continue.
Packets Routinely Lost If packets are routinely lost, proceed to the troubleshooting section in this chapter relating to your network.
Problems Transferring or Copying Files If you have problems copying files out of or into the instrument, you might be experiencing timeout problems. See the previous section on “Timeout Errors.”
Common Problems After a Connection is Made
This section describes common problems you may encounter when using the instrument on a LAN. It assumes you have been able to connect to the instrument in the past. If this is not so, refer to the previous sections first.
Cannot connect to the analyzer
— Make sure the instrument is turned on.
— Check the physical connection to the LAN.
— Make sure the internet (IP) Address of the instrument is set up
correctly in the LAN port setup menu. (Press System, Config I/O, IP Address.)
— When connecting to your instrument over a closed network
(directly through a hub or crossover cable) it may help to set the instrument to its default settings for subnet mask and gateway. (subnet mask: 255.255.0.0, gateway 0.0.0.0)
— If the instrument and the computer are on different networks or
subnets, make sure the gateway address and subnet mask values
Programming Fundamentals · Programming Fundamentals

Using the LAN to Control the Instrument
are set correctly.
Cannot access the file system via ftp
— If the power to the instrument was just turned on, make sure that
you wait about 25 seconds before attempting the connection.
— Verify the LAN connection between your computer and the
instrument. Refer to "If you cannot connect to the instrument" earlier in this section.
Cannot telnet to the command parser port
— Check the telnet port number from the front panel keys.
— Verify the LAN connection between your computer and the
instrument. Refer to "If you cannot connect to the instrument" earlier in this section.
— If the instrument was just turned on, make sure that you wait
about 25 seconds before attempting the connection.
An "operation timed-out" message
Cannot access internal web pages or import graphic images when using a point-to-point connection
If all else fails
Programming Fundamentals · Programming Fundamentals

Using the LAN to Control the Instrument
Pinging the Instrument from a Computer or Workstation
Verify the communications link between the computer and the instrument remote file server using the ping utility.
From a UNIX workstation, type:
ping hostname 64 10
where 64 is the packet size, and 10 is the number of packets transmitted.
From a DOS or Windows environment, type:
where 10 is the number of echo requests.
Normal Response for UNIX
A normal response to the ping will be a total of 9, 10, or possibly 11 packets received with a minimal average round-trip time. The minimal average will be different from network to network. LAN traffic will cause the round-trip time to vary widely.
Because the number of packets received depends on your network traffic and integrity, the normal number might be different for your network.
Normal Response for DOS or Windows
A normal response to the ping will be a total of 9, 10, or possibly 11 packets received if 10 echo requests were specified.
Because the number of packets received depends on your network traffic and integrity, the normal number might be different for your network.
If error messages appear, then check the command syntax before continuing with the troubleshooting. If the syntax is correct, then resolve the error messages using your network documentation, or by consulting your network administrator.
If an unknown host error message appears, then check that the host name and IP address for your instrument are correctly entered from the front panel. Press System, Config I/O.
No Response No packets received indicates no response from a ping.
If there is no response, try typing in the IP address with the ping command, instead of using the host name. Check that the typed address matches the IP address assigned in the System, Config I/O menu, then check the other addresses in the menu.
Check that the host name and IP address are correctly entered in

Using the LAN to Control the Instrument
the node names database.
If you are using a UNIX environment, ping each node along the route between your workstation and the instrument, starting with the your workstation. Ping each gateway, then attempt a ping of the remote file server.
If the instrument still does not respond to ping, then you should suspect a hardware problem with the instrument. To check the instrument performance, refer to "Verify the Instrument Performance" in this chapter.
Intermittent Response If you received 1 to 8 packets back, there is probably a problem with the network. Because the number of packets received depends on your network traffic and integrity, the number might be different for your network.
Use a LAN analyzer or LAN management software to monitor activity and determine where bottlenecks or other problems are occurring. The instrument will still function, but communications over the LAN will be slower.
On a single-client/single-server network, the most likely cause of intermittent response to an echo request is a hardware problem with the LAN module installed in the PC, the cable, or the instrument. To check the instrument, refer to "Verify the Instrument Performance" later in this chapter.
The Standard UNIX PING Command Synopsis ping [-r] [-v] [-o] host [packetsize] [count]
Description The ping command sends an echo request packet to the host once per second. Each echo response packet that is returned is listed on the screen, along with the round-trip time of the echo request and echo response.
Options and Parameters -r Bypasses the routing tables, and sends the
request directly to the host.
-v Reports all packets that are received, including the response packets.
-o Requests information about the network paths taken by the requests and responses.
host The host name or IP address.
packetsize The size of each packet (8 bytes - 4096 bytes).
count The number of packets to send before ending ping (1-(231-1)). If count is not specified, ping sends packets until interrupted.
Programming Fundamentals · Programming Fundamentals

Using the LAN to Control the Instrument
EIA/TIA 568B Wiring Information
Table 6-3 Straight-Through Cable (Unshielded-twisted-pair (UTP) cable with RJ-45 connectors)
Standard, Straight-Through Wiring (each end)
Signal Name RJ-45 Pin # Wire Color Pair #
RX+ 1 white/orange 2
TX+ 3 white/green 3
Table 6-4 Cross-Over Cable (Unshielded-twisted-pair (UTP) cable with RJ-45
Connector A Connector B
Signal Name RJ-45 Pin # RJ-45 Pin # Signal Name
RX2 6 TX-
TX+ 3 1 RX+
TX6 2 RX-
4 4 Not Used 5 5
a. Either end of this cable can be used at the instrument or
LAN device. The connector names are a convention use- ful during cable construction only.
This cable can be used to cascade hubs or to make point-to-point connections without a LAN hub.

Using the LAN to Control the Instrument
NOTE A convenient way to make a cross-over adapter is to use two RJ-45 jacks wired according to Table 6-4, above. Standard straight-through patch cables can then be used from the instrument to the adapter, and from the adapter to other LAN devices. If you use a special-purpose adapter, you will avoid having a cross-over cable mistaken for a standard, straight-through patch cable.
Figure 6-2 Cross-Over Patch Cable Wiring (cross-over end)

Programming Fundamentals · Programming Fundamentals

Programming in C Using the VTL
The programming examples that are provided are written using the C programming language and the Agilent VTL (VISA transition library). This section includes some basic information about programming in the C language. Note that some of this information may not be relevant to your particular application. (For example, if you are not using VXI instruments, the VXI references will not be relevant).
Refer to your C programming language documentation for more details. (This information is taken from the manual “VISA Transition Library”, part number E2090-90026.) The following topics are included:
“Typical Example Program Contents” on page 546 “Linking to VTL Libraries” on page 547 “Compiling and Linking a VTL Program” on page 547 “Example Program” on page 549 “Including the VISA Declarations File” on page 549 “Opening a Session” on page 549 “Device Sessions” on page 550 “Addressing a Session” on page 551 “Closing a Session” on page 553
The following is a summary of the VTL function calls used in the example programs.
visa.h This file is included at the beginning of the file to provide the function prototypes and constants defined by VTL.
ViSession The ViSession is a VTL data type. Each object that will establish a communication channel must be defined as ViSession.
viOpenDefaultRM You must first open a session with the default
resource manager with the viOpenDefaultRM function. This function will initialize the default resource manager and return a pointer to that resource manager session.
viOpen This function establishes a communication channel with the device specified. A session identifier that can be used with other VTL functions is returned. This call must be made for each device you will be using.
viPrintf viScanf These are the VTL formatted I/O functions that are patterned after those used in the C programming
Programming Fundamentals · Programming Fundamentals

Programming in C Using the VTL
language. The viPrintf call sends the IEEE 488.2 *RST command to the instrument and puts it in a known state. The viPrintf call is used again to query for the device identification (*IDN?). The viScanf call is then used to read the results.
viClose This function must be used to close each session. When you close a device session, all data structures that had been allocated for the session will be de-allocated. When you close the default manager session, all sessions opened using the default manager session will be closed.
Your application must link to one of the VTL import libraries:
32-bit Version:
C:\VXIPNP\WIN95\LIB\MSC\VISA32.LIB for Microsoft compilers
C:\VXIPNP\WIN95\LIB\BC\VISA32.LIB for Borland compilers
16-bit Version:
C:\VXIPNP\WIN\LIB\MSC\VISA.LIB for Microsoft compilers
C:\VXIPNP\WIN\LIB\BC\VISA.LIB for Borland compilers
See the following section, “Compiling and Linking a VTL Program” for information on how to use the VTL run-time libraries.
The following is a summary of important compiler-specific considerations for several C/C++ compiler products when developing WIN32 applications.
For Microsoft Visual C++ version 2.0 compilers:
Programming Fundamentals · 32-bit Applications · Programming Fundamentals

Programming in C Using the VTL
They are set by doing the following:
For Borland C++ version 4.0 compilers:
The following is a summary of important compiler-specific considerations for the Windows compiler.
For Microsoft Visual C++ version 1.5:
Otherwise, the library and include files should be explicitly specified in the project file.
Programming Fundamentals · 16-bit Applications · Programming Fundamentals

Programming in C Using the VTL
This example program queries a GPIB device for an identification string and prints the results. Note that you must change the address.
/*idn.c - program filename */
#include "visa.h" #include <stdio.h>
void main () {
/*Open session to GPIB device at address 18 */ ViOpenDefaultRM (&defaultRM); ViOpen (defaultRM, GPIB0::18::INSTR", VI_NULL,
/*Initialize device */ viPrintf (vi, "*RST\n");
/*Send an *IDN? string to the device */ printf (vi, "*IDN?\n");
/*Read results */ viScanf (vi, "%t", &buf);
/*Print results */ printf ("Instrument identification string: %s\n", buf);
/* Close sessions */ viClose (vi); viClose (defaultRM); }
For C and C++ programs, you must include the visa.h header file at the beginning of every file that contains VTL function calls:
#include "visa.h"
This header file contains the VISA function prototypes and the definitions for all VISA constants and error codes. The visa.h header file includes the visatype.h header file.
The visatype.h header file defines most of the VISA types. The VISA types are used throughout VTL to specify data types used in the functions. For example, the viOpenDefaultRM function requires a pointer to a parameter of type ViSession. If you find ViSession in the visatype.h header file, you will find that ViSession is eventually typed as an unsigned long.
A session is a channel of communication. Sessions must first be opened
Programming Fundamentals · VI_NULL, &vi); · Programming Fundamentals

Programming in C Using the VTL
on the default resource manager, and then for each device you will be using. The following is a summary of sessions that can be opened:
NOTE All devices that you will be using need to be connected and in working condition prior to the first VTL function call (viOpenDefaultRM). The system is configured only on the first viOpenDefaultRM per process. Therefore, if viOpenDefaultRM is called without devices connected and then called again when devices are connected, the devices will not be recognized. You must close ALL resource manager sessions and re-open with all devices connected and in working condition.
There are two parts to opening a communications session with a specific device. First you must open a session to the default resource manager with the viOpenDefaultRM function. The first call to this function initializes the default resource manager and returns a session to that resource manager session. You only need to open the default manager session once. However, subsequent calls to viOpenDefaultRM returns a session to a unique session to the same default resource manager resource.
Next, you open a session with a specific device with the viOpen function. This function uses the session returned from viOpenDefaultRM and returns its own session to identify the device session. The following shows the function syntax:
viOpenDefaultRM (sesn);
viOpen (sesn, rsrcName, accessMode, timeout, vi);
The session returned from viOpenDefaultRM must be used in the sesn parameter of the viOpen function. The viOpen function then uses that session and the device address specified in the rsrcName parameter to open a device session. The vi parameter in viOpen returns a session identifier that can be used with other VTL functions.
Your program may have several sessions open at the same time by
Programming Fundamentals · Programming Fundamentals

Programming in C Using the VTL
creating multiple session identifiers by calling the viOpen function multiple times.
The following summarizes the parameters in the previous function calls:
sesn This is a session returned from the viOpenDefaultRM function that identifies the resource manager session.
rsrcName This is a unique symbolic name of the device (device address).
accessMode This parameter is not used for VTL. Use VI_NULL.
timeout This parameter is not used for VTL. Use VI_NULL.
vi This is a pointer to the session identifier for this particular device session. This pointer will be used to identify this device session when using other VTL functions.
The following is an example of opening sessions with a GPIB multimeter and a GPIB-VXI scanner:
ViSession defaultRM, dmm, scanner; . . viOpenDefaultRM(&defaultRM); viOpen (defaultRM, "GPIB0::22::INSTR", VI_NULL,
VI_NULL, &dmm); viOpen (defaultRM, "GPIB-VXI0::24::INSTR", VI_NULL,
VI_NULL, &scanner); . . viClose (scanner); viClose (dmm); viClose(defaultRM);
The above function first opens a session with the default resource manager. The session returned from the resource manager and a device address is then used to open a session with the GPIB device at address 22. That session will now be identified as dmm when using other VTL functions. The session returned from the resource manager is then used again with another device address to open a session with the GPIB-VXI device at primary address 9 and VXI logical address 24. That session will now be identified as scanner when using other VTL functions. See the following section for information on addressing particular devices.
As seen in the previous section, the rsrcName parameter in the viOpen function is used to identify a specific device. This parameter is made up of the VTL interface name and the device address. The interface name is determined when you run the VTL Configuration Utility. This name
Programming Fundamentals · Programming Fundamentals

Programming in C Using the VTL
is usually the interface type followed by a number. The following table illustrates the format of the rsrcName for the different interface types:
VXI VXI [board]::VXI logical address[::INSTR]
GPIB-VXI GPIB-VXI [board]::VXI logical address[::INSTR]
GPIB GPIB [board]::primary address[::secondary address][::INSTR]
The following describes the parameters used above:
board This optional parameter is used if you have more than one interface of the same type. The default value for board is 0.
VSI logical address This is the logical address of the VXI instrument.
primary address This is the primary address of the GPIB device.
secondary address This optional parameter is the secondary address of the GPIB device. If no secondary address is specified, none is assumed.
INSTR This is an optional parameter that indicates that you are communicating with a resource that is of type INSTR, meaning instrument.
The following are examples of valid symbolic names:
XI0::24::INSTR Device at VXI logical address 24 that is of VISA type
INSTR.
VXI2::128 Device at VXI logical address 128, in the third VXI system (VXI2).
GPIB-VXI0::24 A VXI device at logical address 24. This VXI device is
connected via a GPIB-VXI command module.
GPIB0::7::0 A GPIB device at primary address 7 and secondary address 0 on the GPIB interface.
The following is an example of opening a device session with the GPIB device at primary address23.

Programming in C Using the VTL
viOpen (defaultRM, "GPIB0::23::INSTR", VI_NULL,VI_NULL,&vi);
The viClose function must be used to close each session. You can close the specific device session, which will free all data structures that had been allocated for the session. If you close the default resource manager session, all sessions opened using that resource manager will be closed.
Since system resources are also used when searching for resources (viFindRsrc) or waiting for events (viWaitOnEvent), the viClose function needs to be called to free up find lists and event contexts.

Overview of the GPIB Bus
An instrument that is part of a GPIB network is categorized as a listener, talker, or controller, depending on its current function in the network.
Listener A listener is a device capable of receiving data or commands from other instruments. Any number of instruments in the GPIB network can be listeners simultaneously.
Talker A talker is a device capable of transmitting data or commands to other instruments. To avoid confusion, a GPIB system allows only one device at a time to be an active talker.
Controller A controller is an instrument, typically a computer, capable of managing the various GPIB activities. Only one device at a time can be an active controller.
Command statements form the nucleus of GPIB programming. They are understood by all instruments in the network. When combined with the programming language codes, they provide all management and data communication instructions for the system. Refer to the your programming language manual and your computers I/O programming manual for more information.
The seven fundamental command functions are as follows:
Programming Fundamentals · Programming Fundamentals

Overview of the GPIB Bus
Programming Fundamentals · Programming Fundamentals

Overview of the GPIB Bus
Programming Fundamentals · Programming Fundamentals

Using the STATus System

Using the STATus System
Using the STATus System
When you are programming the instrument you may need to monitor instrument status to check for error conditions or monitor changes. You can determine the state of certain instrument events/conditions by programming the status register system. IEEE common commands (those beginning with *) access the higher-level summary registers. To access the information from specific registers you would use the STATus commands.
This chapter includes:

Using the STATus System
Using the STATus System
*ESE <number>, *ESE? Standard Event Status Enable
*ESR? Standard Event Status Register Query
*OPC Operation Complete Command
*OPC? Operation Complete Query
*OPT? Option Information Query
*RCL <register> Recall Instrument State
*RST Reset the Instrument
*PSC, *PSC? Power-on Status Complete
*SAV <register> Save Instrument State
*SRE <number>, *SRE? Service Request Enable
*STB? Read Status Byte Query
*TRG Trigger a Sweep/Measurement
Operation Registers STATus:OPERation:<keyword>
Preset the Status Bytes STATus:PRESet
Status Registers STATus:QUEStionable:<keyword>

Using the STATus System
Using the STATus System
Calibration Registers STATus:QUEStionable:CALibration:<keyword>
Questionable Calibration Event Query
Questionable Calibration Negative Transition
Questionable Calibration Positive Transition
Frequency Registers STATus:QUEStionable:FREQuency:<keyword>
Questionable Frequency Event Query
Questionable Frequency Negative Transition
Questionable Frequency Positive Transition
Integrity Registers STATus:QUEStionable:INTegrity:<keyword>
Questionable Integrity Event Query
Questionable Integrity Event Query
Questionable Integrity Negative Transition
Questionable Integrity Positive Transition
Signal Integrity Registers STATus:QUEStionable:INTegrity:SIGNal:<keyword>
Questionable Integrity Signal Condition
Questionable Integrity Signal Enable
Questionable Integrity Signal Event Query
Questionable Integrity Signal Negative Transition
Questionable Integrity Signal Positive Transition
STATus:QUEStionable:INTegrity:UNCalibrated:<keyw ord>

Using the STATus System
Using the STATus System
Questionable Calibration Integrity Condition
Questionable Calibration Integrity Enable
Questionable Calibration Integrity Event Query
Questionable Calibration Integrity Negative Transition
Questionable Calibration Integrity Positive Transition
Power Registers STATus:QUEStionable:POWer:<keyword>
Questionable Power Event Query
Questionable Power Negative Transition
Questionable Power Positive Transition
Temperature Registers STATus:QUEStionable:TEMPerature:<keyword>
Questionable Temperature Event Query
Questionable Temperature Negative Transition
Questionable Temperature Positive Transition

Using the STATus System
Using the STATus System
Status System Commands

Using the STATus System
Using the STATus System
These commands are specified in IEEE Standard 488.2-1992, IEEE Standard Codes, Formats, Protocols and Common Commands for Use with ANSI/IEEE Std 488.1-1987. New York, NY, 1992.
Numeric values for bit patterns can be entered using decimal or hexidecimal representations. (i.e. 0 to 32767 is equivalent to #H0 to #H7FFF) See the SCPI Basics information about using bit patterns for variable parameters.
*CAL?
Performs a full alignment and returns a number indicating the success of the alignment. A zero is returned if the alignment is successful. A one is returned if any part of the alignment fails. The equivalent SCPI command is CALibrate[:ALL]?
*CLS
Clears the status byte. It does this by emptying the error queue and clearing all bits in all of the event registers. The status byte registers summarize the states of the other registers. It is also responsible for generating service requests.
Key Type: There is no equivalent front panel key.
*ESE?
Selects the desired bits from the standard event status enable register. This register monitors I/O errors and synchronization conditions such as operation complete, request control, query error, device dependent error, execution error, command error and power on. The selected bits are OR’d to become a summary bit (bit 5) in the status byte register which can be queried.
The query returns the state of the standard event status enable
Common IEEE Commands · *ESE

Using the STATus System
Using the STATus System
register.
Key Type: There is no equivalent front panel key.
*ESR?
Queries and clears the standard event status event register. (This is a destructive read.)
Key Type: There is no equivalent front panel key.
*IDN?
Returns an instrument identification information. The string will contain the model number, serial number and firmware revision.
The response is organized into four fields separated by commas. The field definitions are as follows:
For example:
Agilent Technologies,E4440A,US00000123,B.02.02 (for PSA)
Key Type: There is no equivalent front panel key.
Remarks: An @ in the firmware revision information indicates that it is proto firmware.
*LRN?
This command is not implemented. Other commands are available for
Common IEEE Commands

Using the STATus System
Using the STATus System
returning the *LRN data. Use *IDN? to return the instrument model number, serial number, and firmware version. Use the *SAV/*RCL commands to save and then return the instrument state information.
*OPC
*OPC?
The *OPC command sets bit 0 in the standard event status register to “1” when pending operations have finished. It does not hold off subsequent operations.
The *OPC? query stops new commands from being processed until the current processing is complete. Then it returns a “1”, and the program continues. This query can be used to synchronize events of other instruments on the external bus.
The instrument does not wait for completion of all processes for these commands. The processes that are monitored are identified in the STATus:OPERation register. These include:
PSA Process STATus:OPER
Register Bit Byte Value
MEASuring (not in all modes)
Waiting for trigger 5 32
For example, if you want to verify the completion of both calibration and waiting for trigger set :STAT:OPER:ENAB 35 and monitor any changes.
Key Type: There is no equivalent front panel key.
*OPT?
For ESA and PSA analyzers: This command is not implemented. However the information is available remotely by using SCPI commands. See :SYSTem:OPTions? in the SYSTem subsystem.
Common IEEE Commands · Calibrating 0 1 · Sweeping 3 8

Using the STATus System
Using the STATus System
*PSC
*PSC?
Sets the state of the power-on status clear flag. This command allows you to specify if the service request enable register and the event status enable register should be cleared at power up.
The query returns the contents of the power-on status clear flag.
Key Type: There is no equivalent front panel key.
This command recalls the instrument state from the specified instrument memory register.
If the state being loaded has a newer firmware revision than the revision of the instrument, no state is recalled and an error is reported.
If the state being loaded has an equal firmware revision than the revision of the instrument, the state will be loaded.
If the state being loaded has an older firmware revision than the revision of the instrument, the instrument will only load the parts of the state that apply to the older revision.
Key Type: There is no equivalent front panel key.
:MMEMory:STORe:STATe
*RST
This command presets the instrument to a factory defined condition that is appropriate for remote programming operation. In Spectrum Analysis Mode *RST is equivalent to performing the commands
Common IEEE Commands · *RCL

Using the STATus System
Using the STATus System
*RST does not change the mode and only resets the parameters for the current mode.
The :SYSTem:PRESet command is equivalent to a front panel Preset key.
This command saves the instrument state to the specified instrument memory register.
Key Type: There is no equivalent front panel key.
*SRE?
This command sets the value of the service request enable register.
The query returns the value of the register.
Key Type: There is no equivalent front panel key.
*STB?
Returns the value of the status byte register without erasing its contents.
Key Type: There is no equivalent front panel key.
Common IEEE Commands · *SAV · *SRE

Using the STATus System
Using the STATus System
*TRG
This command triggers the instrument. Use the :TRIGger[:SEQuence]:SOURce command to select the trigger source.
If you have selected a one-button measurement and it has been paused (INITiate:PAUSe), or the CONFigure: command was used. The command causes the system to exit this “waiting” state and go to the “initiated” state. The trigger system is initiated and completes one full trigger cycle. It returns to the “waiting” state on completion of the trigger cycle. See the MEASURE key for more information about controlling the measurement process.
The instrument must be in the single measurement mode. If INIT:CONT ON, then the command is ignored. Depending upon the measurement and the number of averages, there may be multiple data acquisitions, with multiple trigger events, for one full trigger cycle.
Key Type: There is no equivalent front panel key.
*TST?
For PSA analyzers, NO tests are performed. *TST? always returns 0.
Key Type: There is no equivalent front panel key.
*WAI
This command causes the instrument to wait until all pending commands/processes are completed before executing any additional commands. There is no query form for the command.
The instrument does not wait for completion of all processes. The processes that are monitored are identified in the *OPC? command description.
Key Type: There is no equivalent front panel key.
Common IEEE Commands

Using the STATus System
Using the STATus System
Common IEEE Commands

Using the STATus System
Using the STATus System
Common IEEE Commands

Using the STATus System
Using the STATus System
The STATus subsystem controls the SCPI-defined instrument-status reporting structures. Each status register has a set of five commands used for querying or masking that particular register.
Numeric values for bit patterns can be entered using decimal or hexidecimal representations. (i.e. 0 to 32767 is equivalent to #H0 to #H7FFF) See the SCPI Basics information about using bit patterns for variable parameters.
:STATus:OPERation:CONDition?
This query returns the decimal value of the sum of the bits in the Status Operation Condition register.
Key Type: There is no equivalent front panel key.
:STATus:OPERation:ENABle
:STATus:OPERation:ENABle?
This command determines which bits in the Operation Event register, will set the Operation Status Summary bit (bit 7) in the Status Byte Register. The variable is the sum of the decimal values of the bits you want to enable.
Key Type: There is no equivalent front panel key.
STATus Subsystem · Operation Condition Query · Operation Enable

Using the STATus System
Using the STATus System
:STATus:OPERation[:EVENt]?
This query returns the decimal value of the sum of the bits in the Operation Event register.
The data in this register is latched until it is queried. Once queried, the register is cleared.
Key Type: There is no equivalent front panel key.
:STATus:OPERation:NTRansition
:STATus:OPERation:NTRansition?
This command determines which bits in the Operation Condition register will set the corresponding bit in the Operation Event register when the condition register bit has a negative transition (1 to 0). The variable is the sum of the decimal values of the bits that you want to enable.
Key Type: There is no equivalent front panel key.
:STATus:OPERation:PTRansition
:STATus:OPERation:PTRansition?
This command determines which bits in the Operation Condition register will set the corresponding bit in the Operation Event register when the condition register bit has a positive transition (0 to 1). The variable is the sum of the decimal values of the bits that you want to enable.
Key Type: There is no equivalent front panel key.

Using the STATus System
Using the STATus System
:STATus:PRESet
Sets bits in most of the enable and transition registers to their default state. It presets all the Transition Filters, Enable Registers, and the Error/Event Queue Enable. It has no effect on Event Registers, Error/Event QUEue, IEEE 488.2 ESE, and SRE Registers as described in IEEE Standard 488.2-1992, IEEE Standard Codes, Formats, Protocols and Common Commands for Use with ANSI/IEEE Std 488.1-1987. New York, NY, 1992.
Key Type: There is no equivalent front panel key.
:STATus:QUEStionable:CONDition?
This query returns the decimal value of the sum of the bits in the Questionable Condition register.
Key Type: There is no equivalent front panel key.
:STATus:QUEStionable:ENABle
:STATus:QUEStionable:ENABle?
This command determines which bits in the Questionable Event register will set the Questionable Status Summary bit (bit3) in the Status Byte Register. The variable is the sum of the decimal values of the bits you want to enable.
NOTE The preset condition is all bits in this enable register set to 0. To have any Questionable Events reported to the Status Byte Register, one or more bits need to be set to 1. The Status Byte Event Register should be queried after each measurement to check the Questionable Status Summary (bit 3). If it is equal to 1, a condition during the test may have made the test results invalid. If it is equal to 0, this indicates that no hardware problem or measurement problem was detected by the analyzer.
STATus Subsystem · Questionable Condition · Questionable Enable

Using the STATus System
Using the STATus System
Key Type: There is no equivalent front panel key.
:STATus:QUEStionable[:EVENt]?
This query returns the decimal value of the sum of the bits in the Questionable Event register.
The data in this register is latched until it is queried. Once queried, the register is cleared.
Key Type: There is no equivalent front panel key.
:STATus:QUEStionable:NTRansition
:STATus:QUEStionable:NTRansition?
This command determines which bits in the Questionable Condition register will set the corresponding bit in the Questionable Event register when the condition register bit has a negative transition (1 to 0). The variable is the sum of the decimal values of the bits that you want to enable.
Key Type: There is no equivalent front panel key.
:STATus:QUEStionable:PTRansition
:STATus:QUEStionable:PTRansition?
This command determines which bits in the Questionable Condition register will set the corresponding bit in the Questionable Event register when the condition register bit has a positive transition (0 to 1). The variable is the sum of the decimal values of the bits that you want to enable.

Using the STATus System
Using the STATus System
Key Type: There is no equivalent front panel key.
:STATus:QUEStionable:CALibration:CONDition?
This query returns the decimal value of the sum of the bits in the Questionable Calibration Condition register.
Key Type: There is no equivalent front panel key.
:STATus:QUEStionable:CALibration:ENABle
:STATus:QUEStionable:CALibration:ENABle?
This command determines which bits in the Questionable Calibration Condition Register will set bits in the Questionable Calibration Event register, which also sets the Calibration Summary bit (bit 8) in the Questionable Register. The variable is the sum of the decimal values of the bits you want to enable.
Key Type: There is no equivalent front panel key.
Example STAT:QUES:CAL:ENABLE 16384 could be used if you have turned off the automatic alignment and you want to query if an alignment is needed.
Questionable Calibration Event Query
:STATus:QUEStionable:CALibration[:EVENt]?
This query returns the decimal value of the sum of the bits in the Questionable Calibration Event register.
STATus Subsystem · Questionable Calibration Condition · Questionable Calibration Enable

Using the STATus System
Using the STATus System
The data in this register is latched until it is queried. Once queried, the register is cleared.
Key Type: There is no equivalent front panel key.
Questionable Calibration Negative Transition
:STATus:QUEStionable:CALibration:NTRansition
:STATus:QUEStionable:CALibration:NTRansition?
This command determines which bits in the Questionable Calibration Condition register will set the corresponding bit in the Questionable Calibration Event register when the condition register bit has a negative transition (1 to 0). The variable is the sum of the decimal values of the bits that you want to enable.
Key Type: There is no equivalent front panel key.
Questionable Calibration Positive Transition
:STATus:QUEStionable:CALibration:PTRansition
:STATus:QUEStionable:CALibration:PTRansition?
This command determines which bits in the Questionable Calibration Condition register will set the corresponding bit in the Questionable Calibration Event register when the condition register bit has a positive transition (0 to 1). The variable is the sum of the decimal values of the bits that you want to enable.
Key Type: There is no equivalent front panel key.
:STATus:QUEStionable:FREQuency:CONDition?
STATus Subsystem · Questionable Frequency Condition

Using the STATus System
Using the STATus System
This query returns the decimal value of the sum of the bits in the Questionable Frequency Condition register.
Key Type: There is no equivalent front panel key.
:STATus:QUEStionable:FREQuency:ENABle
:STATus:QUEStionable:FREQuency:ENABle?
This command determines which bits in the Questionable Frequency Condition Register will set bits in the Questionable Frequency Event register, which also sets the Frequency Summary bit (bit 5) in the Questionable Register. The variable is the sum of the decimal values of the bits you want to enable.
Key Type: There is no equivalent front panel key.
Questionable Frequency Event Query
:STATus:QUEStionable:FREQuency[:EVENt]?
This query returns the decimal value of the sum of the bits in the Questionable Frequency Event register.
The data in this register is latched until it is queried. Once queried, the register is cleared.
Key Type: There is no equivalent front panel key.
Questionable Frequency Negative Transition
:STATus:QUEStionable:FREQuency:NTRansition
:STATus:QUEStionable:FREQuency:NTRansition?
This command determines which bits in the Questionable Frequency
STATus Subsystem · Questionable Frequency Enable

Using the STATus System
Using the STATus System
Condition register will set the corresponding bit in the Questionable Frequency Event register when the condition register bit has a negative transition (1 to 0). The variable is the sum of the decimal values of the bits that you want to enable.
Key Type: There is no equivalent front panel key.
Questionable Frequency Positive Transition
:STATus:QUEStionable:FREQuency:PTRansition
:STATus:QUEStionable:FREQuency:PTRansition?
This command determines which bits in the Questionable Frequency Condition register will set the corresponding bit in the Questionable Frequency Event register when the condition register bit has a positive transition (0 to 1). The variable is the sum of the decimal values of the bits that you want to enable.
Key Type: There is no equivalent front panel key.
:STATus:QUEStionable:INTegrity:CONDition?
This query returns the decimal value of the sum of the bits in the Questionable Integrity Condition register.
Key Type: There is no equivalent front panel key.
:STATus:QUEStionable:INTegrity:ENABle
:STATus:QUEStionable:INTegrity:ENABle?
This command determines which bits in the Questionable Integrity
STATus Subsystem · Questionable Integrity Condition · Questionable Integrity Enable

Using the STATus System
Using the STATus System
Condition Register will set bits in the Questionable Integrity Event register, which also sets the Integrity Summary bit (bit 9) in the Questionable Register. The variable is the sum of the decimal values of the bits you want to enable.
Key Type: There is no equivalent front panel key.
Questionable Integrity Event Query
:STATus:QUEStionable:INTegrity[:EVENt]?
This query returns the decimal value of the sum of the bits in the Questionable Integrity Event register.
The data in this register is latched until it is queried. Once queried, the register is cleared.
Key Type: There is no equivalent front panel key.
Questionable Integrity Negative Transition
:STATus:QUEStionable:INTegrity:NTRansition
:STATus:QUEStionable:INTegrity:NTRansition?
This command determines which bits in the Questionable Integrity Condition register will set the corresponding bit in the Questionable Integrity Event register when the condition register bit has a negative transition (1 to 0) The variable is the sum of the decimal values of the bits that you want to enable.
Key Type: There is no equivalent front panel key.
Questionable Integrity Positive Transition
:STATus:QUEStionable:INTegrity:PTRansition
:STATus:QUEStionable:INTegrity:PTRansition?
STATus Subsystem

Using the STATus System
Using the STATus System
This command determines which bits in the Questionable Integrity Condition register will set the corresponding bit in the Questionable Integrity Event register when the condition register bit has a positive transition (0 to 1). The variable is the sum of the decimal values of the bits that you want to enable.
Key Type: There is no equivalent front panel key.
Questionable Integrity Signal Condition
:STATus:QUEStionable:INTegrity:SIGNal:CONDition?
This query returns the decimal value of the sum of the bits in the Questionable Integrity Signal Condition register.
Key Type: There is no equivalent front panel key.
Questionable Integrity Signal Enable
:STATus:QUEStionable:INTegrity:SIGNal:ENABle
:STATus:QUEStionable:INTegrity:SIGNal:ENABle?
This command determines which bits in the Questionable Integrity Signal Condition Register will set bits in the Questionable Integrity Signal Event register, which also sets the Integrity Summary bit (bit 9) in the Questionable Register. The variable is the sum of the decimal values of the bits you want to enable.
Key Type: There is no equivalent front panel key.
Questionable Integrity Signal Event Query
:STATus:QUEStionable:INTegrity:SIGNal[:EVENt]?
This query returns the decimal value of the sum of the bits in the
STATus Subsystem

Using the STATus System
Using the STATus System
Questionable Integrity Signal Event register.
The data in this register is latched until it is queried. Once queried, the register is cleared.
Key Type: There is no equivalent front panel key.
Questionable Integrity Signal Negative Transition
:STATus:QUEStionable:INTegrity:SIGNal:NTRansition
:STATus:QUEStionable:INTegrity:SIGNal:NTRansition?
This command determines which bits in the Questionable Integrity Signal Condition register will set the corresponding bit in the Questionable Integrity Signal Event register when the condition register bit has a negative transition (1 to 0). The variable is the sum of the decimal values of the bits that you want to enable.
Key Type: There is no equivalent front panel key.
Questionable Integrity Signal Positive Transition
:STATus:QUEStionable:INTegrity:SIGNal:PTRansition
:STATus:QUEStionable:INTegrity:SIGNal:PTRansition?
This command determines which bits in the Questionable Integrity Signal Condition register will set the corresponding bit in the Questionable Integrity Signal Event register when the condition register bit has a positive transition (0 to 1). The variable is the sum of the decimal values of the bits that you want to enable.
Key Type: There is no equivalent front panel key.
STATus Subsystem

Using the STATus System
Using the STATus System
Questionable Integrity Uncalibrated Condition
:STATus:QUEStionable:INTegrity:UNCalibrated:CONDition?
This query returns the decimal value of the sum of the bits in the Questionable Integrity Uncalibrated Condition register.
Key Type: There is no equivalent front panel key.
Questionable Integrity Uncalibrated Enable
:STATus:QUEStionable:INTegrity:UNCalibrated:ENABle
:STATus:QUEStionable:INTegrity:UNCalibrated:ENABle?
This command determines which bits in the Questionable Integrity Uncalibrated Condition Register will set bits in the Questionable Integrity Uncalibrated Event register, which also sets the Data Uncalibrated Summary bit (bit 3) in the Questionable Integrity Register. The variable is the sum of the decimal values of the bits you want to enable.
Key Type: There is no equivalent front panel key.
Questionable Integrity Uncalibrated Event Query
:STATus:QUEStionable:INTegrity:UNCalibrated[:EVENt]?
This query returns the decimal value of the sum of the bits in the Questionable Integrity Uncalibrated Event register.
The data in this register is latched until it is queried. Once queried, the register is cleared.
Key Type: There is no equivalent front panel key.
STATus Subsystem

Using the STATus System
Using the STATus System
Questionable Integrity Uncalibrated Negative Transition
:STATus:QUEStionable:INTegrity:UNCalibrated:NTRansition
:STATus:QUEStionable:INTegrity:UNCalibrated:NTRansition?
This command determines which bits in the Questionable Integrity Uncalibrated Condition register will set the corresponding bit in the Questionable Integrity Uncalibrated Event register when the condition register bit has a negative transition (1 to 0). The variable is the sum of the decimal values of the bits that you want to enable.
Key Type: There is no equivalent front panel key.
Questionable Integrity Uncalibrated Positive Transition
:STATus:QUEStionable:INTegrity:UNCalibrated:PTRansition
:STATus:QUEStionable:INTegrity:UNCalibrated:PTRansition?
This command determines which bits in the Questionable Integrity Uncalibrated Condition register will set the corresponding bit in the Questionable Integrity Uncalibrated Event register when the condition register bit has a positive transition (0 to 1). The variable is the sum of the decimal values of the bits that you want to enable.
Key Type: There is no equivalent front panel key.
:STATus:QUEStionable:POWer:CONDition?
This query returns the decimal value of the sum of the bits in the Questionable Power Condition register.
Key Type: There is no equivalent front panel key.
STATus Subsystem · Questionable Power Condition

Using the STATus System
Using the STATus System
:STATus:QUEStionable:POWer:ENABle
:STATus:QUEStionable:POWer:ENABle?
This command determines which bits in the Questionable Power Condition Register will set bits in the Questionable Power Event register, which also sets the Power Summary bit (bit 3) in the Questionable Register. The variable is the sum of the decimal values of the bits you want to enable.
Key Type: There is no equivalent front panel key.
Questionable Power Event Query
:STATus:QUEStionable:POWer[:EVENt]?
This query returns the decimal value of the sum of the bits in the Questionable Power Event register.
The data in this register is latched until it is queried. Once queried, the register is cleared.
Key Type: There is no equivalent front panel key.
Questionable Power Negative Transition
:STATus:QUEStionable:POWer:NTRansition
:STATus:QUEStionable:POWer:NTRansition?
This command determines which bits in the Questionable Power Condition register will set the corresponding bit in the Questionable Power Event register when the condition register bit has a negative transition (1 to 0). The variable is the sum of the decimal values of the bits that you want to enable.
Key Type: There is no equivalent front panel key.
STATus Subsystem · Questionable Power Enable

Using the STATus System
Using the STATus System
Questionable Power Positive Transition
:STATus:QUEStionable:POWer:PTRansition
:STATus:QUEStionable:POWer:PTRansition?>
This command determines which bits in the Questionable Power Condition register will set the corresponding bit in the Questionable Power Event register when the condition register bit has a positive transition (0 to 1). The variable is the sum of the decimal values of the bits that you want to enable.
Key Type: There is no equivalent front panel key.
:STATus:QUEStionable:TEMPerature:CONDition?
This query returns the decimal value of the sum of the bits in the Questionable Temperature Condition register.
Key Type: There is no equivalent front panel key.
:STATus:QUEStionable:TEMPerature:ENABle
:STATus:QUEStionable:TEMPerature:ENABle?
This command determines which bits in the Questionable Temperature Condition Register will set bits in the Questionable Temperature Event register, which also sets the Temperature Summary bit (bit 4) in the Questionable Register. The variable is the sum of the decimal values of the bits you want to enable.
Key Type: There is no equivalent front panel key.
STATus Subsystem · Questionable Temperature Condition · Questionable Temperature Enable

Using the STATus System
Using the STATus System
Questionable Temperature Event Query
:STATus:QUEStionable:TEMPerature[:EVENt]?
This query returns the decimal value of the sum of the bits in the Questionable Temperature Event register.
The data in this register is latched until it is queried. Once queried, the register is cleared
Key Type: There is no equivalent front panel key.
Questionable Temperature Negative Transition
:STATus:QUEStionable:TEMPerature:NTRansition
:STATus:QUEStionable:TEMPerature:NTRansition?
This command determines which bits in the Questionable Temperature Condition register will set the corresponding bit in the Questionable Temperature Event register when the condition register bit has a negative transition (1 to 0). The variable is the sum of the decimal values of the bits that you want to enable.
Key Type: There is no equivalent front panel key.
Questionable Temperature Positive Transition
:STATus:QUEStionable:TEMPerature:PTRansition
:STATus:QUEStionable:TEMPerature:PTRansition?
This command determines which bits in the Questionable Temperature Condition register will set the corresponding bit in the Questionable Temperature Event register when the condition register bit has a positive transition (0 to 1). The variable is the sum of the decimal values of the bits that you want to enable.
Key Type: There is no equivalent front panel key.
STATus Subsystem

This chapter provide a visual representation of the front-panel keys and their associated menu keys (Spectrum Analysis mode is selected). These menu maps are in alphabetical order by the front-panel key label or oval cross-reference label. You can locate detailed information about each key/function at the page number listed in the figure title for each menu. For one-button measurement menus see Chapter 9 on page 623.

Alpha Editor Keys, 1 of 2


Alpha Editor Keys, 1 of 2

Alpha Editor Keys, 2 of 2

AMPLITUDE Y Scale Key, 1 of 2 (See page 49)
‡ A double-dagger to the left of the softkey indicates a function that is not always available. It is dependent on other instrument settings.

A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.
† A dagger to the left of the softkey indicates that when the key is pressed this is an active function.

AMPLITUDE Y Scale Key, 1 of 2 (See page 49)

† A dagger to the left of the softkey indicates that when the key is pressed this is an active function.
AMPLITUDE Y Scale Key, 2 of 2 (See page 49)

Auto Couple Key, 1 of 3 (See page 71)
‡ A double-dagger to the left of the softkey indicates a function that is not always available. It is dependent on other instrument settings.
A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.
† A dagger to the left of the softkey indicates that when the key is pressed this is an active function.


Auto Couple Key, 1 of 3 (See page 71)
‡ A double-dagger to the left of the softkey indicates a function that is not always available. It is dependent on other instrument settings.

A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.
Auto Couple Key, 2 of 3 (See page 71)

Auto Couple Key, 1 of 3 (See page 71)
A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.

Auto Couple Key, 3 of 3 (See page 71)

BW/Avg Key (See page 83)

A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.
† A dagger to the left of the softkey indicates that when the key is pressed this is an active function.

Det/Demod Key (See page 95)
A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.


Display Key, 1 of 2 (See page 103)

† A dagger to the left of the softkey indicates that when the key is pressed this is an active function.

Display Key, 1 of 2 (See page 103)

† A dagger to the left of the softkey indicates that when the key is pressed this is an active function.
Display Key, 2 of 2 (See page 103)

File Key, 1 of 6 (See page 127)


File Key, 1 of 6 (See page 127)
‡ A double-dagger to the left of the softkey indicates a function that is not always available. It is dependent on other instrument settings.

File Key, 2 of 6 (See page 127)

File Key, 1 of 6 (See page 127)

‡ A double-dagger to the left of the softkey indicates a function that is not always available. It is dependent on other instrument settings.
File Key, 3 of 6 (See page 127)

File Key, 1 of 6 (See page 127)

File Key, 4 of 6 (See page 127)

File Key, 1 of 6 (See page 127)

A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.
File Key, 5 of 6 (See page 127)

File Key, 1 of 6 (See page 127)

A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.
File Key, 6 of 6 (See page 127)

FREQUENCY Channel Key (See page 163)
† A dagger to the left of the softkey indicates that when the key is pressed this is an active function.


Input/Output Key (See page 171)
A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.


Marker Key (See page 177)
‡ A double-dagger to the left of the softkey indicates a function that is not always available. It is dependent on other instrument settings.

A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.
† A dagger to the left of the softkey indicates that when the key is pressed this is an active function.

Marker --> Key (See page 197)
‡ A double-dagger to the left of the softkey indicates a function that is not always available. It is dependent on other instrument settings.


Marker Fctn Key (See page 191)

A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.
† A dagger to the left of the softkey indicates that when the key is pressed this is an active function.

MODE Key (See page 203)


Peak Search Key (See page 209)


Preset Key (See page 221)


Print Setup Key (See page 229)

‡ A double-dagger to the left of the softkey indicates a function that is not always available. It is dependent on other instrument settings.
A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.

SPAN X Scale Key (See page 237)


SPAN X Scale Key for CCDF Measurement (See page 389)


Sweep Key (See page 241)


System Key, 1 of 4 (See page 247)


System Key, 1 of 4 (See page 247)

A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.
† A dagger to the left of the softkey indicates that when the key is pressed this is an active function.
System Key, 2 of 4 (See page 247)

System Key, 1 of 4 (See page 247)

System Key, 3 of 4 (See page 247)

System Key, 1 of 4 (See page 247)
‡ A double-dagger to the left of the softkey indicates a function that is not always available. It is dependent on other instrument settings.

System Key, 4 of 4 (See page 247)

Trace/View Key (See page 283)
A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.


Trig Key (See page 287)
‡ A double-dagger to the left of the softkey indicates a function that is not always available. It is dependent on other instrument settings.
A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.


This chapter provides you with menu maps for all measurements available to you when you press the front-panel Measure key (Spectrum Analysis mode is selected).

One-Button Measurement Menu Maps
This section provides a visual representation of all front-panel key menus associated with the measurements available when you press the front-panel Measure key. The key menus common to all measurements (Measure, Meas Control, and Mode Setup) are presented first. Then, the key menus unique to each measurement are displayed. Use the following table to locate the menus you wish to view and Table on page 294 to locate the key descriptions and programming commands for these measurements.
Menu Map Location Menu Map Location
Measure Page 625 Mult-Carrier Power Menus
Meas Control Page 626 Meas Setup Page 643
Mode Setup Page 627 Trace/View Page 644
ACP Menus Occupied BW Menus
Meas Setup Page 629 Meas Setup Page 645
Trace/View Page 630 Spectrum Emission Mask Menus
Burst Power Menus Display Page 648
Meas Setup Page 631 Marker Page 650
Trace/View Page 632 Meas Setup Page 646
CCDF Menus (Power Stat) Span X Scale Page 649
Display Page 635 Trace/View Page 647
Marker Page 637 Spurious Emissions Menus
Meas Setup Page 633 Meas Setup Page 651
Span X Scale Page 636
Meas Setup Page 638
Meas Setup Page 640
Meas Setup Page 642

One-Button Measurement Menu Maps


One-Button Measurement Menu Maps

Functions

One-Button Measurement Menu Maps

A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.

One-Button Measurement Menu Maps

A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.
Functions

One-Button Measurement Menu Maps

A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.
† A dagger to the left of the softkey indicates that when the key is pressed this is an active function.

One-Button Measurement Menu Maps

A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.
Functions

One-Button Measurement Menu Maps

† A dagger to the left of the softkey indicates that when the key is pressed this is an active function.

One-Button Measurement Menu Maps

A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.
Functions

One-Button Measurement Menu Maps

† A dagger to the left of the softkey indicates that when the key is pressed this is an active function.

One-Button Measurement Menu Maps

A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.
Functions

One-Button Measurement Menu Maps


One-Button Measurement Menu Maps
Functions

One-Button Measurement Menu Maps

A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.
† A dagger to the left of the softkey indicates that when the key is pressed this is an active function.

One-Button Measurement Menu Maps

† A dagger to the left of the softkey indicates that when the key is pressed this is an active function.
Functions

One-Button Measurement Menu Maps

A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.

One-Button Measurement Menu Maps
‡ A double-dagger to the left of the softkey indicates a function that is not always available. It is dependent on other instrument settings.

† A dagger to the left of the softkey indicates that when the key is pressed this is an active function.
Functions

One-Button Measurement Menu Maps
A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.


One-Button Measurement Menu Maps

† A dagger to the left of the softkey indicates that when the key is pressed this is an active function.
Functions

One-Button Measurement Menu Maps

‡ A double-dagger to the left of the softkey indicates a function that is not always available. It is dependent on other instrument settings.
† A dagger to the left of the softkey indicates that when the key is pressed this is an active function.

One-Button Measurement Menu Maps

A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.
Functions

One-Button Measurement Menu Maps

† A dagger to the left of the softkey indicates that when the key is pressed this is an active function.

One-Button Measurement Menu Maps

A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.
† A dagger to the left of the softkey indicates that when the key is pressed this is an active function.
Functions

One-Button Measurement Menu Maps
A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.


One-Button Measurement Menu Maps

Functions

One-Button Measurement Menu Maps


One-Button Measurement Menu Maps

A bar on the left of two or more softkeys indicates that the keys are a set of mutually exclusive choices.
Functions

One-Button Measurement Menu Maps

† A dagger to the left of the softkey indicates that when the key is pressed this is an active function.

One-Button Measurement Menu Maps
Functions











Index · Index - 12