Agilent B2200 User’s Guidep.11 Introductionp.17Agilent B2200 Seriesp.19Front Panelp.20Rear Panelp.22Switch Modulesp.24Specificationsp.26General Specificationsp.27Switch Modulesp.28Supplemental Information for B2200A/B2210Ap.29Supplemental Information for B2201A/B2211Ap.30Accessories and Optionsp.312 Installationp.35Requirementsp.37Power Requirementsp.37Power Cablep.37Operating Environmentp.39Storage and Shipping Environmentp.39Inspectionp.40Installing the B2200p.41To Set the GPIB Addressp.42To Connect the GPIB Cablep.42To Install the Switch Modulep.43To Install the Blank Panelp.43Self-Testp.44Output Connectionsp.45Output Connectorsp.45Connector Platesp.46To Make Connections to DUT Interfacep.47To Make Interlock Circuitp.49To Mount Connectorsp.52Input Connectionsp.54Measurement Cable Lengthp.57Maintenancep.59Calibrationp.59Cleaningp.593 Front Panel Operationp.61Operationp.63To Initialize Agilent B2200p.64To Enable Light Penp.64To Change Channel Configuration Modep.64To Change Connection Rulep.65To Change Connection Sequencep.65To Control Switch Conditionp.65To Open All Switchesp.66To Save/Load Setup Datap.66To Use Bias Modep.67To Use Ground Modep.68To Use Couple Modep.69To Display Firmware Revisionp.69To Display Module Informationp.69To Read Error Messagep.70To Set Beeperp.70To Set GPIB Addressp.70To Set Remote Display Modep.71To Return to Local Modep.71Switch Control Functionsp.72Channel Configuration Modep.73Connection Rulep.74Connection Sequencep.75Bias Modep.76Ground Modep.78Couple Modep.80Display Functionsp.82LED Matrixp.82LCDp.83Front Panel Keysp.85Function key groupp.86Edit key groupp.89Setup Menusp.90Setup Menup.90Selftest Menup.944 Programmingp.99Programming Basicsp.101SCPI Command Hierarchyp.101Fundamental Commandsp.102Switch Controlp.103Programming Examplesp.105Connecting Input-Output Pathsp.106Using Bias Modep.108Using Ground Modep.110Using Couple Modep.112Saving Input/Output Labelsp.114Defining Comment for Internal Memoryp.116Capacitance Compensationp.118Capacitance Compensation Functionp.118Required Conditionsp.119To Create Compensation Data Filep.121To Perform Measurement and Compensationp.1255 SCPI Command Referencep.127Common Commandsp.130Commands Summaryp.130*CLSp.131*ESEp.131*ESR?p.132*IDN?p.133*OPCp.134*RSTp.135*SREp.136*STB?p.137*TST?p.138*WAIp.138Subsystem Commandsp.139Command Summaryp.139:DIAGnostic:TEST:CARD:CLEarp.147:DIAGnostic:TEST:CARD[:EXECute]?p.147:DIAGnostic:TEST:CARD:STATe?p.148:DIAGnostic:TEST:FRAMe:CLEarp.148:DIAGnostic:TEST:FRAMe[:EXECute]?p.149:DIAGnostic:TEST:FRAMe:STATe?p.150[:ROUTe]:AGND:CHANnel:DISable:CARDp.150[:ROUTe]:AGND:CHANnel:DISable[:LIST]p.151[:ROUTe]:AGND:CHANnel:ENABle:CARDp.151[:ROUTe]:AGND:CHANnel:ENABle[:LIST]p.152[:ROUTe]:AGND:PORTp.153[:ROUTe]:AGND[:STATe]p.154[:ROUTe]:AGND:UNUSEDp.155[:ROUTe]:BIAS:CHANnel:DISable:CARDp.156[:ROUTe]:BIAS:CHANnel:DISable[:LIST]p.156[:ROUTe]:BIAS:CHANnel:ENABle:CARDp.157[:ROUTe]:BIAS:CHANnel:ENABle[:LIST]p.157[:ROUTe]:BIAS:PORTp.158[:ROUTe]:BIAS[:STATe]p.159[:ROUTe]:CLOSe:CARD?p.160[:ROUTe]:CLOSe[:LIST]p.160[:ROUTe]:CONNection:RULEp.161[:ROUTe]:CONNection:SEQuencep.162[:ROUTe]:COUPle:PORTp.163[:ROUTe]:COUPle:PORT:DETectp.164[:ROUTe]:COUPle[:STATe]p.165[:ROUTe]:FUNCtionp.166[:ROUTe]:OPEN:CARDp.166[:ROUTe]:OPEN[:LIST]p.167[:ROUTe]:SYMBol:CHANnelp.168[:ROUTe]:SYMBol:PORTp.169:SYSTem:BEEPp.170:SYSTem:CCONfig?p.170:SYSTem:CDEScription?p.171:SYSTem:CPONp.172:SYSTem:CTYPe?p.173:SYSTem:DISPlay:LCDp.173:SYSTem:DISPlay:LEDp.174:SYSTem:DISPlay:STRingp.174:SYSTem:ERRor?p.175:SYSTem:KLCp.175:SYSTem:MEMOry:COMMentp.176:SYSTem:MEMOry:DELetep.176:SYSTem:MEMOry:LOADp.177:SYSTem:MEMOry:SAVEp.177:SYSTem:PENp.178:SYSTem:VERSion?p.178Status Reporting Structurep.179Status Reporting Structurep.179Status Byte Registerp.181Service Request Enable Registerp.183Standard Event Status Registerp.184Standard Event Status Enable Registerp.185Output Queuep.1866 VXI plug&play Driverp.187System Requirementsp.189Installing VXIplug&play Driverp.190Driver Functionsp.191agb220xa_biasChanCardp.193agb220xa_biasChanListp.193agb220xa_biasChanList_Qp.194agb220xa_biasPortp.195agb220xa_biasStatep.195agb220xa_closep.196agb220xa_closeCard_Qp.196agb220xa_closeListp.197agb220xa_closeList_Qp.198agb220xa_cmdp.198agb220xa_cmdData_Qp.199agb220xa_cmdIntp.199agb220xa_cmdInt16Arr_Qp.200agb220xa_cmdInt16_Qp.200agb220xa_cmdInt32Arr_Qp.201agb220xa_cmdInt32_Qp.201agb220xa_cmdRealp.202agb220xa_cmdReal64Arr_Qp.202agb220xa_cmdReal64_Qp.203agb220xa_cmdString_Qp.203agb220xa_compenCp.204agb220xa_connRuleSeqp.204agb220xa_couplePortp.205agb220xa_coupleStatep.206agb220xa_dclp.206agb220xa_detectCouplePortp.206agb220xa_error_messagep.207agb220xa_error_queryp.207agb220xa_errorQueryDetectp.208agb220xa_errorQueryDetect_Qp.208agb220xa_esr_Qp.209agb220xa_funcp.209agb220xa_groundChanCardp.210agb220xa_groundChanListp.210agb220xa_groundChanList_Qp.211agb220xa_groundPortp.212agb220xa_groundStatep.212agb220xa_initp.213agb220xa_opc_Qp.213agb220xa_openCardp.214agb220xa_openListp.214agb220xa_openList_Qp.215agb220xa_readStatusByte_Qp.215agb220xa_resetp.216agb220xa_revision_queryp.216agb220xa_selectCompenFilep.216agb220xa_self_testp.217agb220xa_testClearp.217agb220xa_testExec_Qp.218agb220xa_timeOutp.218agb220xa_timeOut_Qp.219agb220xa_unusedPortp.2197 Error Messagesp.221Standard SCPI Error Messagesp.223Command Errorp.223Execution Errorp.227Device-Dependent Errorsp.228Query Errorsp.229B2200 Specific Error Messagesp.230B2200 Channel Related Errorsp.231B2200 Card/Mode/Port Related Errorsp.232
Page 1 of the Agilent Technologies B2201A user manual — Agilent Technologies
Page 1 · click to enlarge
Page 1
Agilent Technologies B2201A — Agilent B2200 User’s Guide diagram (page 1)
Fig. 1 · Agilent B2200 User’s Guide

Agilent B2200A Femto Leakage Switch Mainframe Agilent B2201A 14ch Low Leakage Switch Mainframe

User’s Guide

Agilent Technologies

Page 2 of the Agilent Technologies B2201A user manual — Restricted Rights Legend
Page 2 · click to enlarge
Page 2

Notices

© Agilent Technologies 2004, 2005

Warranty

defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.

No part of this manual may be reproduced in any form or by any means (including elec- tronic storage and retrieval or translation into a foreign language) without prior agree- ment and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.

The material contained in this docu- ment is provided “as is,” and is sub- ject to being changed, without notice, in future editions. Further, to the max- imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par- ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connec- tion with the furnishing, use, or per- formance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep- arate agreement shall control.

Manual Part Number

B2200-90000

Edition

Edition 1, October 2004 Edition 2, July 2005

Agilent Technologies, Inc. 5301 Stevens Creek Blvd Santa Clara, CA 95051 USA

Technology Licenses

The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor- dance with the terms of such license.

Restricted Rights Legend

If software is for use in the performance of a U.S. Government prime contract or subcon- tract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer soft- ware” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclo- sure of Software is subject to Agilent Tech- nologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as

Page 3 of the Agilent Technologies B2201A user manual — Agilent B2200 User’s Guide
Page 3 · click to enlarge
Page 3

DECLARATION OF CONFORMITY According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014

Manufacturer’s NameAgilent Technologies International sarl Manufacturer’s Address: Supplier’s Address:

Rue de la Gare 29 CH - 1110 Morges Switzerland

Declares under sole responsibility that the product as originally delivered

Product NameFemto Leakage Switch Mainframe

Femto Leakage Switch Module 14ch Low Leakage Switch Mainframe 14ch Low Leakage Switch Module Model Number: Agilent B2200A

Agilent B2210A Agilent B2201A

Agilent B2211A Product OptionsThis declaration covers all options of the above product(s)

complies with the essential requirements of the following applicable European Directives, and carries the CE marking accordingly:

Low Voltage Directive (73/23/EEC, amended by 93/68/EEC) EMC Directive (89/336/EEC, amended by 93/68/EEC) and conforms with the following product standards EMC Standard IEC 61326:2002 / EN 61326:1997 +A1:1998 +A2:2001 +A3:2003 CISPR 11:1997 / EN 55011:1998

Group 1 Class A 4 kV CD, 8 kV AD 3 V/m, 80-1000 MHz 0.5 kV signal lines, 1 kV power lines 0.5 kV line-line, 1 kV line-ground 3 V, 0.15-80 MHz 1 cycle, 100%

IEC61000-4-2:1995 / EN61000-4-2:1995 IEC 61000-4-3:1995 / EN61000-4-3:1995 IEC 61000-4-4:1995 / EN61000-4-4:1995 IEC 61000-4-5:1995 / EN61000-4-5:1995 IEC 61000-4-6:1996 / EN61000-4-6:1996 IEC 61000-4-11:1994 / EN61000-4-11:1994

CanadaICES-001:1998 Australia/New Zealand: AS/NZS 2064.1

The product was tested in a typical configuration with Agilent Technologies test systems.

Safety IEC 61010-1:2001 / EN 61010-1:2001 Canada: CSA C22.2 No. 1010.1:1992, NRTL/C

Supplementary InformationThis DoC applies to above-listed products placed on the EU market after:

QA Manager Agilent Technologies

Limit · September 26, 2006 · Date Toshiyuki Kawaji

Page 4 of the Agilent Technologies B2201A user manual — ACOUSTIC NOISE EMISSION
Page 4 · click to enlarge
Page 4
  • Herstellerbescheinigung

GEÄUSCHEMISSION

Lpa < 70 dB

nach DIN 45635 T. 19

  • Manufacturer’s Declaration

ACOUSTIC NOISE EMISSION

Lpa < 70 dB

Cet appareil ISM est conforme ?Hla norme NMB-001 du Canada.

This product complies with the WEEE Directive (2002/96/EC) marking requirements. The affixed label indicates that you must not discard this electrical/ electronic product in domestic household waste.

Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as a “Monitoring and Control instrumentation” product.

Do not dispose in domestic household waste.

To return unwanted products, contact your local Agilent office, or see www.agilent.com/environment/product/ for more information.

Microsoft, Windows, Windows NT, Visual C++, and Visual Basic are registered trademarks of Microsoft Corporation.

Borland C++ Builder is registered trademark of International, Inc. LabWindows and LabVIEW are registered

trademarks of National Instruments Corporation. All other trademarks are the property of their respective owners.

Show 5 more labels / short text on this page
am Arbeitsplatz · normaler Betrieb · operator position · normal operation · per ISO 7779
Page 5 of the Agilent Technologies B2201A user manual — Safety Summary
Page 5 · click to enlarge
Page 5

Safety Summary

The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual may impair the protections provided by the equipment. In addition, it violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies, Inc. assumes no liability for customer’s failure to comply with these requirements.

Agilent B2200 is INDOOR USE products.

  • GROUND THE INSTRUMENT

This is Safety Class I instrument. To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical ground. The power terminal and the power cable must meet International Electrotechnical Commission (IEC) safety standards.

  • DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE

Do not operate the instrument in the presence of flammable gases or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard.

  • KEEP AWAY FROM LIVE CIRCUITS

Operation personnel must not remove instrument covers. Component replacement and internal adjustments must be made by qualified maintenance personnel. Do not replace components with power cable connected. Under certain conditions, dangerous voltages may exist even with the power cable removed. To avoid injuries, always disconnect power and discharge circuits before touching them.

  • DO NOT SERVICE OR ADJUST ALONE

Do not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present.

Page 6 of the Agilent Technologies B2201A user manual — Agilent B2200 User’s Guide
Page 6 · click to enlarge
Page 6
  • DO NOT SUBSTITUTE PARTS OR MODIFY INSTRUMENT

Because of the danger of introducing additional hazards, do not install substitute parts or perform any unauthorized modification to the instrument. Return the instrument to a Agilent Technologies Sales and Service Office for services and repair to ensure that safety features are maintained.

  • DANGEROUS PROCEDURE WARNINGS
Page 7 of the Agilent Technologies B2201A user manual — CAT 1
Page 7 · click to enlarge
Page 7

Safety Symbols

The general definitions of safety symbols used on equipment or in manuals are listed below.

Instruction manual symbol: the product will be marked with this symbol when it is necessary for the user to refer to the instruction manual in order to protect against damage to the instrument.

Indicates dangerous voltage and potential for electrical shock. Do not touch terminals that have this symbol when insrument is on.

Affixed to product containing static sensitive devices--use anti-static handling procedures to prevent electrostatic discharge damage to component.

Protective conductor terminal. For protection against electrical shock in case of a fault. Used with field wiring terminals to indicate the terminal which must be connected to ground before operating equipment.

Frame or chassis terminal. A connection to the frame (chassis) of the equipment which normally includes all exposed metal structures.

Indicates earth (ground) terminal.

Alternating current.

Direct current.

ON (Supply).

OFF (Supply).

STANDBY (Supply).

CAT 1

Means INSTALLATION CATEGORY I. Measurement terminals on the rear panel comply with INSTALLATION CATEGORY I.

Page 8 of the Agilent Technologies B2201A user manual — In This Manual
Page 8 · click to enlarge
Page 8

In This Manual

This manual is a user’s guide for Agilent B2200A/B2201A Switch Mainframe, and consists of the following chapters:

  • Introduction

Describes an overview and specifications of the Agilent B2200 series.

  • Installation

Describes how to install the Agilent B2200 and how to setup the input/output.

  • Front Panel Operation

Explains the front panel operation and the switch control functions of the Agilent B2200, also provides the reference information of the front panel keys and display.

  • Programming

Explains the programming examples that control the Agilent B2200.

  • SCPI Command Reference

Describes the all commands of the Agilent B2200 SCPI commands.

  • VXI plug&play Driver

Describes the all functions of the Agilent B2200 VXIplug&play driver.

  • Error Messages

Lists and describes the error messages for the Agilent B2200.

The following text conventions are used in this manual:

Screen Text Represents text that appears on screen of the controller.

Italic Refers to a related document, or is used for emphasis.

Text Conventions

Page 9 of the Agilent Technologies B2201A user manual — Contents
Page 9 · click to enlarge
Page 9

Contents

1. Introduction

Agilent B2200 Series 1-3

Front Panel 1-4

1-4 Rear Panel 1-6

Switch Modules 1-8

1-8

Specifications 1-10

General Specifications 1-11 Switch Modules 1-12 Supplemental Information for B2200A/B2210A 1-13 Supplemental Information for B2201A/B2211A 1-14

Accessories and Options 1-15

2. Installation

2-2

Requirements 2-3

Power Requirements 2-3 Power Cable 2-3 Operating Environment 2-5 Storage and Shipping Environment 2-5

Inspection 2-6

Installing the B2200 2-7

To Set the GPIB Address 2-8 To Connect the GPIB Cable 2-8 To Install the Switch Module 2-9

2-9 To Install the Blank Panel 2-9

Page 10 of the Agilent Technologies B2201A user manual — Contents
Page 10 · click to enlarge
Page 10

Contents

Self-Test 2-10

Output Connections 2-11

Output Connectors 2-11 Connector Plates 2-12 To Make Connections to DUT Interface 2-13 To Make Interlock Circuit 2-15 To Mount Connectors 2-18

Input Connections 2-20

Measurement Cable Length 2-23

Maintenance 2-25

Calibration 2-25 Cleaning 2-25

3. Front Panel Operation

Operation 3-3

To Initialize Agilent B2200 3-4 To Enable Light Pen 3-4 To Change Channel Configuration Mode 3-4 To Change Connection Rule 3-5 To Change Connection Sequence 3-5 To Control Switch Condition 3-5 To Open All Switches 3-6 To Save/Load Setup Data 3-6 To Use Bias Mode 3-7 To Use Ground Mode 3-8 To Use Couple Mode 3-9 To Display Firmware Revision 3-9 To Display Module Information 3-9 To Read Error Message 3-10 To Set Beeper 3-10

Page 11 of the Agilent Technologies B2201A user manual — Contents
Page 11 · click to enlarge
Page 11

Contents

To Set GPIB Address 3-10 To Set Remote Display Mode 3-11 To Return to Local Mode 3-11

Switch Control Functions 3-12

Channel Configuration Mode 3-13 Connection Rule 3-14 Connection Sequence 3-15 Bias Mode 3-16 Ground Mode 3-18 Couple Mode 3-20

Display Functions 3-22

LED Matrix 3-22 LCD 3-23

Front Panel Keys 3-25

Function key group 3-26 Edit key group 3-29

Setup Menus 3-30

Setup Menu 3-30 Selftest Menu 3-34

4. Programming

Programming Basics 4-3

SCPI Command Hierarchy 4-3 Fundamental Commands 4-4 Switch Control 4-5

Programming Examples 4-7

Connecting Input-Output Paths 4-8 Using Bias Mode 4-10 Using Ground Mode 4-12

Page 12 of the Agilent Technologies B2201A user manual — Contents
Page 12 · click to enlarge
Page 12

Contents

Using Couple Mode 4-14 Saving Input/Output Labels 4-16 Defining Comment for Internal Memory 4-18

Capacitance Compensation 4-20

Capacitance Compensation Function 4-20 Required Conditions 4-21 To Create Compensation Data File 4-23 To Perform Measurement and Compensation 4-27

5. SCPI Command Reference

Common Commands 5-4

Commands Summary 5-4 *CLS 5-5 *ESE 5-5 *ESR? 5-6 *IDN? 5-7 *OPC 5-8 *RST 5-9 *SRE 5-10 *STB? 5-11 *TST? 5-12 *WAI 5-12

Subsystem Commands 5-13

Command Summary 5-13 :DIAGnostic:TEST:CARD:CLEar 5-21 :DIAGnostic:TEST:CARD[:EXECute]? 5-21 :DIAGnostic:TEST:CARD:STATe? 5-22 :DIAGnostic:TEST:FRAMe:CLEar 5-22 :DIAGnostic:TEST:FRAMe[:EXECute]? 5-23 :DIAGnostic:TEST:FRAMe:STATe? 5-24 [:ROUTe]:AGND:CHANnel:DISable:CARD 5-24

Page 13 of the Agilent Technologies B2201A user manual — Contents
Page 13 · click to enlarge
Page 13

Contents

[:ROUTe]:AGND:CHANnel:DISable[:LIST] 5-25 [:ROUTe]:AGND:CHANnel:ENABle:CARD 5-25 [:ROUTe]:AGND:CHANnel:ENABle[:LIST] 5-26 [:ROUTe]:AGND:PORT 5-27 [:ROUTe]:AGND[:STATe] 5-28 [:ROUTe]:AGND:UNUSED 5-29 [:ROUTe]:BIAS:CHANnel:DISable:CARD 5-30 [:ROUTe]:BIAS:CHANnel:DISable[:LIST] 5-30 [:ROUTe]:BIAS:CHANnel:ENABle:CARD 5-31 [:ROUTe]:BIAS:CHANnel:ENABle[:LIST] 5-31 [:ROUTe]:BIAS:PORT 5-32 [:ROUTe]:BIAS[:STATe] 5-33 [:ROUTe]:CLOSe:CARD? 5-34 [:ROUTe]:CLOSe[:LIST] 5-34 [:ROUTe]:CONNection:RULE 5-35 [:ROUTe]:CONNection:SEQuence 5-36 [:ROUTe]:COUPle:PORT 5-37 [:ROUTe]:COUPle:PORT:DETect 5-38 [:ROUTe]:COUPle[:STATe] 5-39 [:ROUTe]:FUNCtion 5-40 [:ROUTe]:OPEN:CARD 5-40 [:ROUTe]:OPEN[:LIST] 5-41 [:ROUTe]:SYMBol:CHANnel 5-42 [:ROUTe]:SYMBol:PORT 5-43 :SYSTem:BEEP 5-44 :SYSTem:CCONfig? 5-44 :SYSTem:CDEScription? 5-45 :SYSTem:CPON 5-46 :SYSTem:CTYPe? 5-47 :SYSTem:DISPlay:LCD 5-47 :SYSTem:DISPlay:LED 5-48 :SYSTem:DISPlay:STRing 5-48

Page 14 of the Agilent Technologies B2201A user manual — Contents
Page 14 · click to enlarge
Page 14

Contents

:SYSTem:ERRor? 5-49 :SYSTem:KLC 5-49 :SYSTem:MEMOry:COMMent 5-50 :SYSTem:MEMOry:DELete 5-50 :SYSTem:MEMOry:LOAD 5-51 :SYSTem:MEMOry:SAVE 5-51 :SYSTem:PEN 5-52 :SYSTem:VERSion? 5-52

Status Reporting Structure 5-53

Status Reporting Structure 5-53 Status Byte Register 5-55 Service Request Enable Register 5-57 Standard Event Status Register 5-58 Standard Event Status Enable Register 5-59 Output Queue 5-60

6. VXI plug&play Driver

System Requirements 6-3

Installing VXIplug&play Driver 6-4

Driver Functions 6-5

agb220xa_biasChanCard 6-7 agb220xa_biasChanList 6-7 agb220xa_biasChanList_Q 6-8 agb220xa_biasPort 6-9 agb220xa_biasState 6-9 agb220xa_close 6-10 agb220xa_closeCard_Q 6-10 agb220xa_closeList 6-11 agb220xa_closeList_Q 6-12 agb220xa_cmd 6-12

Page 15 of the Agilent Technologies B2201A user manual — Contents
Page 15 · click to enlarge
Page 15

Contents

agb220xa_cmdData_Q 6-13 agb220xa_cmdInt 6-13 agb220xa_cmdInt16Arr_Q 6-14 agb220xa_cmdInt16_Q 6-14 agb220xa_cmdInt32Arr_Q 6-15 agb220xa_cmdInt32_Q 6-15 agb220xa_cmdReal 6-16 agb220xa_cmdReal64Arr_Q 6-16 agb220xa_cmdReal64_Q 6-17 agb220xa_cmdString_Q 6-17 agb220xa_compenC 6-18 agb220xa_connRuleSeq 6-18 agb220xa_couplePort 6-19 agb220xa_coupleState 6-20 agb220xa_dcl 6-20 agb220xa_detectCouplePort 6-20 agb220xa_error_message 6-21 agb220xa_error_query 6-21 agb220xa_errorQueryDetect 6-22 agb220xa_errorQueryDetect_Q 6-22 agb220xa_esr_Q 6-23 agb220xa_func 6-23 agb220xa_groundChanCard 6-24 agb220xa_groundChanList 6-24 agb220xa_groundChanList_Q 6-25 agb220xa_groundPort 6-26 agb220xa_groundState 6-26 agb220xa_init 6-27 agb220xa_opc_Q 6-27 agb220xa_openCard 6-28 agb220xa_openList 6-28 agb220xa_openList_Q 6-29

Page 16 of the Agilent Technologies B2201A user manual — Contents
Page 16 · click to enlarge
Page 16

Contents

agb220xa_readStatusByte_Q 6-29 agb220xa_reset 6-30 agb220xa_revision_query 6-30 agb220xa_selectCompenFile 6-30 agb220xa_self_test 6-31 agb220xa_testClear 6-31 agb220xa_testExec_Q 6-32 agb220xa_timeOut 6-32 agb220xa_timeOut_Q 6-33 agb220xa_unusedPort 6-33

7. Error Messages

Standard SCPI Error Messages 7-3

Command Error 7-3 Execution Error 7-7 Device-Dependent Errors 7-8 Query Errors 7-9

B2200 Specific Error Messages 7-10

B2200 Channel Related Errors 7-11 B2200 Card/Mode/Port Related Errors 7-12

Page 17 of the Agilent Technologies B2201A user manual — 1 Introduction
Page 17 · click to enlarge
Page 17

1 Introduction

Page 18 of the Agilent Technologies B2201A user manual — 1 Introduction
Page 18 · click to enlarge
Page 18

This chapter describes the basic functions and features of the Agilent B2200A/B2201A (Agilent B2200 series), and consists of the following sections:

  • “Agilent B2200 Series”
  • “Front Panel”
  • “Rear Panel”
  • “Switch Modules”
  • “Specifications”
  • “Accessories and Options”

Introduction

Page 19 of the Agilent Technologies B2201A user manual — Agilent B2200 Series
Page 19 · click to enlarge
Page 19

Introduction Agilent B2200 Series

Agilent B2200 Series

Agilent B2200 series is a switching matrix designed for semiconductor dc characteristics measurement applications. The Agilent B2200 series has 14 input ports and four card slots for the switch modules (plug-in cards), and can configure a 12, 24, 36, or 48 outputs switching matrix. The Agilent B2200A/B2201A supports the following dedicated switch module.

  • Agilent B2210A fA Leakage Switch Module

Dedicated for the Agilent B2200A mainframe.

  • Agilent B2211A 14ch Low Leakage Switch Module

Dedicated for the Agilent B2201A mainframe.

Basic functions of the Agilent B2200 series are listed below.

  • SMU inputs (for source monitor units): 8 ports
  • AUX inputs (multipurpose inputs): 6 ports
  • Outputs: 12, 24, 36, or 48 ports
  • Connection setup and status monitor on the front panel
  • Status display on the LED matrix display
  • Connection setup by the Light pen
  • Automatic control through GPIB interface
  • Auto ground function
  • Self-test, relay function test
  • Relay cleaning
Page 20 of the Agilent Technologies B2201A user manual — Front Panel
Page 20 · click to enlarge
Page 20

Front Panel

The Agilent B2200 series provides the front panel keys, the LCD, and the LED matrix display for the status monitor and connection setup.

  • Line switch

Used to turn the Agilent B2200 on or off.

  • LCD

Used to monitor the status and set the connection. See Chapter 3 for the details.

  • Front panel keys

Used to set the Agilent B2200. See Chapter 3 for the details.

  • LED matrix display

Displays the status of the matrix switches. Also used to set the matrix connection with the light pen.

  • SMU Input

Inputs for the source monitor unit (SMU). Eight input ports. Up to four kelvin inputs are available. Triaxial BNC connector.

  • AUX Input

Multipurpose inputs. Six input ports. Coaxial BNC connector. CMH and CML terminals are the input ports for the capacitance measurement.

The input port 12 (AUX Input 12) is the default ground input port. If you assign the port as the ground input port, this ground port will be internally connected to the ground when the ground mode is set to ON. In this case, open this connector.

  • Light Pen

Connector for the Agilent 16443A Light pen. Used to set the matrix connection with the LED matrix display.

Do not apply an input signal over these limits to the input terminals. If you do, the Agilent B2200 will be damaged. If you use a bias source that has current limit capability, set the bias source current limit.

Introduction Front Panel

Page 21 of the Agilent Technologies B2201A user manual — Front Panel
Page 21 · click to enlarge
Page 21

Figure 1-1 Front Panel View

Agilent Technologies B2201A — Front Panel diagram (page 21)
Fig. 2 · Front Panel

Introduction · Front Panel

Page 22 of the Agilent Technologies B2201A user manual — Rear Panel
Page 22 · click to enlarge
Page 22

Rear Panel

The Agilent B2200 series has four card slots for the switch modules, the GPIB interface and so on.

  • Card slots

For the Agilent B2200A mainframe, the Agilent B2210A cards are installed.

For the Agilent B2201A mainframe, the Agilent B2211A cards are installed.

Mixed configuration of the switch modules is not supported.

  • GPIB interface

Use an Agilent 82357A USB/GPIB interface or Agilent 10833A/B/C/D GPIB cable to connect to an external computer or equipment.

  • Serial number

You need this serial number when using the Agilent Technologies telephone assistance program.

  • LINE input receptacle

AC power cable is connected to this receptacle.

Introduction Rear Panel

Page 23 of the Agilent Technologies B2201A user manual — Rear Panel
Page 23 · click to enlarge
Page 23

Figure 1-2 Rear Panel View

Agilent Technologies B2201A — Rear Panel diagram (page 23)
Fig. 3 · Rear Panel

Introduction · Rear Panel

Page 24 of the Agilent Technologies B2201A user manual — Switch Modules
Page 24 · click to enlarge
Page 24
Agilent Technologies B2201A — Switch Modules diagram (page 24)
Fig. 4 · Switch Modules

Switch Modules

The Agilent B2200A and B2201A support the dedicated switch module, Agilent B2210A and B2211A, respectively.

By installing the modules, the module inputs will be connected internally to the front panel input connectors. And 12 output connectors will face the rear panel. The type of the output connectors is the triaxial BNC. Up to six kelvin outputs are available.

Image of the output terminals and the block diagram of the switch module are shown in Figure 1-3 and Figure 1-4.

Figure 1-3 Switch Module Output Terminals

Agilent Technologies B2201A — Switch Modules diagram (page 24)
Fig. 5 · Switch Modules
Page 25 of the Agilent Technologies B2201A user manual — 13 CMH
Page 25 · click to enlarge
Page 25

Figure 1-4 Switch Module Block Diagram

1 2 3 4 5 6 7 8 9 10 11 12

13 CMH

14 CML

Introduction Switch Modules · Output · Input

Page 26 of the Agilent Technologies B2201A user manual — Specifications
Page 26 · click to enlarge
Page 26

Specifications

This section lists specifications and supplemental information for the Agilent B2200 series.

  • “General Specifications”
  • “Switch Modules”
  • “Supplemental Information for B2200A/B2210A”
  • “Supplemental Information for B2201A/B2211A”

The specifications are the performance standards or limits against which these units have been tested. The supplemental information is not warranted, but provides useful information about functions and performance.

If not noted otherwise, the conditions for specifications and supplemental information are as follows:

Temperature23 ° C ± 5 ° C

Humidity: 5 % to 60 % R.H.

Introduction Specifications

Page 27 of the Agilent Technologies B2201A user manual — General Specifications
Page 27 · click to enlarge
Page 27

General Specifications

Temperature range:

Operating5 ° C to 35 ° C
Storage− 20 ° C to 70 ° C
Humidity rangeOperating: 5 % to 70 % R.H., non-condensing

Storage:

< 80 % R.H. at 35 °C, < 60 % R.H. at 65 °C, non-condensing (B2200A/B2210A)

< 80 % R.H. at 65 °C, non-condensing (B2201A/B2211A)

AltitudeOperating: 0 to 2,000 m (6,500 ft)
Storage0 to 15,240 m (50,000 ft)

Regulatory compliance:

SafetyCSA C22.2 No. 1010.1/IEC 1010-1
EMCCISPR 11 Group 1 class A&EN50082-1

Power requirement:

90 to 264 V (continuous), 47 to 63 Hz, 2 A/200 VA maximum

Number of slots4 slots for 48 mm height switch module
DimensionsB2200A/B2201A: 430 mm W × 320 mm H × 600 mm D
B2210A/B2211A395 mm W × 48 mm H × 500 mm D
WeightB2200A/B2201A: approximately 14.0 kg
B2210Aapproximately 5.0 kg
B2211Aapproximately 3.5 kg
Number of portsSMU input: 8 triaxial ports (with Guard)
AUX input6 coaxial ports (two ports for capacitance measurement)
Output12 triaxial ports (with Guard), maximum 48 ports

Introduction Specifications

Page 28 of the Agilent Technologies B2201A user manual — B2210A B2211A
Page 28 · click to enlarge
Page 28

Switch Modules

Table 1-1 lists the specifications when the switch module is installed in the mainframe. In the table, the conditions are as follows:

Conditions: 23 °C ± 5 °C, 5 % to 60 % R.H.

Table 1-1 Agilent B2210A/B2211A Switch Module Specifications

B2210A B2211A

Max. current rating (A)

SMU input 1.0 1.0

AUX input 0.5 0.5

Max. voltage rating (V)

SMU input, channel to common 200 200

SMU input, channel to channel 300 300

AUX input, channel to common 100 100

AUX input, channel to channel 100 100

Closed channel residual resistance (Ω)

SMU input 0.6 0.6

AUX input 1.5 1.5

Channel isolation resistance (Ω)

SMU input 1014 5 × 1013

AUX input 109 109

The Agilent B2200A mainframe supports the Agilent B2210A module.

The Agilent B2201A mainframe supports the Agilent B2211A module.

Mixed configuration of the switch modules is not supported.

Introduction Specifications

Page 29 of the Agilent Technologies B2201A user manual — Supplemental Information for B2200A/B2210A
Page 29 · click to enlarge
Page 29

Supplemental Information for B2200A/B2210A

Offset current 1 < 10 fA (typical 3 fA) (SMU input)

IM noise (RMS) 2 0.6 fA (SMU input)

Channel crosstalk capacitance < 1 pF/channel (SMU input)

< 3 pF/channel (AUX input)

Offset voltage < 50 µV (SMU input)

< 80 µV (AUX input)

Settling time 3 2.0 seconds

Bandwidth (at -3dB) 30 MHz (SMU input)

Guard capacitance 4 < 145 pF (SMU input)

Additional C measurement error 5 < ± 1 % + 0.2 pF (SMU input)

1. When the voltage applied to all input-output channels is 0 V. 2. Measured by the Agilent 4156C with the integration time setting 100 PLC. When the voltage applied to all other input-output paths is 0 V. 3. The time until the measurement value settles to within 50 fA of the final value. When the applied voltage is 10 V. 4. When four modules have been installed in the mainframe. And when only one input-output path is made. 5. For the measurement less than 1000 pF at the frequency 1 kHz to 1 MHz, using the Agilent 4284A with 3 m cable. After the compensation using the capacitance compensation algorithm (a function of VXIplug&play driver).

Introduction Specifications

Page 30 of the Agilent Technologies B2201A user manual — Supplemental Information for B2201A/B2211A
Page 30 · click to enlarge
Page 30

Supplemental Information for B2201A/B2211A

Offset current 1 < 50 fA (SMU input)

IM noise (RMS) 2 5 fA (SMU input)

Channel crosstalk capacitance < 0.5 pF/channel (SMU input)

< 3 pF/channel (AUX input)

Offset voltage < 80 µV (SMU input)

< 100 µV (AUX input)

Settling time 3 2.0 seconds

Bandwidth (at -3dB) 30 MHz (SMU input)

Guard capacitance 4 < 145 pF (SMU input)

Additional C measurement error 5 < ± 1 % + 0.2 pF (SMU input)

1. When the voltage applied to all input-output paths is 0 V. 2. Measured by the Agilent 4156C with the integration time setting 100 PLC. When the voltage applied to all other input-output paths is 0 V. 3. The time until the measurement value settles to within 300 fA of the final value. When the applied voltage is 10 V. 4. When four modules have been installed in the mainframe. And when only one input-output path is made. 5. For the measurement less than 1000 pF at the frequency 1 kHz to 1 MHz, using the Agilent 4284A with 3 m cable. After the compensation using the capacitance compensation algorithm (a function of VXIplug&play driver).

Introduction Specifications

Page 31 of the Agilent Technologies B2201A user manual — Accessories and Options
Page 31 · click to enlarge
Page 31

Introduction Accessories and Options

Accessories and Options

Agilent B2200 is furnished with the following accessories.

  • Power cable, 1 ea.
  • Operation summary sheet, 1 ea.
  • Manual CD-ROM, 1 ea.
  • Software CD-ROM, 1 ea.

Stores the Agilent B2200 VXIplug&play driver.

The Agilent B2200 VXIplug&play driver supports Windows XP Professional, Windows 2000, Windows NT 4.0, Windows 98, and Windows 95.

  • Moisture-proof and dehumidifying packing kit, 1 set

Table 1-2 lists the options and accessories available for the Agilent B2200.

Page 32 of the Agilent Technologies B2201A user manual — Accessories and Options
Page 32 · click to enlarge
Page 32

Introduction Accessories and Options

Table 1-2 Options and Accessories

Model Number Option Item Description

B2200A fA Leakage Switch Mainframe

B2200A-UK6 Commercial cal. certificate w/ test data

B2200A-ABA Manual set, English

B2200A-ABJ Manual set, Japanese

B2201A 14ch Low Leakage Switch Mainframe

B2201A-UK6 Commercial cal. certificate w/ test data

B2201A-ABA Manual set, English

B2201A-ABJ Manual set, Japanese

B2210A fA Leakage Switch Module (for B2200A)

B2211A 14ch Low Leakage Switch Module (for B2201A)

16493H GNDU cable (between 41501/4142 and 16495F/G)

16493H-001 1.5 m length

16493H-002 3 m length

16493J Interlock cable (between E5260/E5270/4155/4156 and 16495F/G)

16493J-001 1.5 m length

16493J-002 3 m length

16493J-003 5 m length

16493K Kelvin triaxial cable (between B2200 inputs and E5260/E5270/4156/41501)

16493K-001 1.5 m length

16493K-002 3 m length

16493L GNDU cable (between E5260/E5270 and 16495F/G)

16493L-001 1.5 m length

16493L-002 3 m length

16493L-003 5 m length

16493N GNDU cable for Kelvin connection (between B2200 inputs and E5260/E5270/41501/4142)

16443A Light pen

Page 33 of the Agilent Technologies B2201A user manual — Accessories and Options
Page 33 · click to enlarge
Page 33

Introduction Accessories and Options

Model Number Option Item Description

16494A-001 1.5 m length

16494A-002 3 m length

16494A-003 80 cm length

16494A-005 4 m length

16494B Kelvin triaxial cable (between B2200 inputs and 4142B, between B2210/B2211 outputs and 16495F/G)

16494B-001 1.5 m length

16494B-002 3 m length

16494B-003 80 cm length

16494C Kelvin triaxial cable (between B2210/B2211 outputs and B2220A)

16494C-001 1.5 m length

16494C-002 3 m length

16494C-005 4 m length

16494F CMU Input cable, 2 m (between B2200 CMH/CML and 4-terminal pairs connectors)

16495E Half size blank plate

16495F Half size connector plate with 12×triaxial, intlk, GNDU

16495F-001 Bulkhead feedthrough connectors (female to female)

16495F-002 Connectors to contacts for soldering

16495G Full size connector plate with 24×triaxial, intlk, GNDU

16495G-001 Bulkhead feedthrough connectors (female to female)

16495G-002 Connectors to contacts for soldering

16494A Triaxial cable

Page 34 of the Agilent Technologies B2201A user manual — Accessories and Options
Page 34 · click to enlarge
Page 34

Introduction Accessories and Options

Page 35 of the Agilent Technologies B2201A user manual — 2 Installation
Page 35 · click to enlarge
Page 35

2 Installation

Page 36 of the Agilent Technologies B2201A user manual — 2 Installation
Page 36 · click to enlarge
Page 36

This chapter describes requirements to install Agilent B2200 and the tasks for installation, and is organized into the following three sections:

  • “Requirements”
  • “Inspection”
  • “Installing the B2200”
  • “Self-Test”
  • “Output Connections”
  • “Input Connections”
  • “Measurement Cable Length”
  • “Maintenance”
  • Use a three-conductor ac power cable to connect cabinet (if used) and the Agilent B2200 to an electric ground (safety ground).
  • If you need to touch the force and guard terminals of the output connector, turn off the Agilent B2200 and discharge any capacitors whenever possible.
  • Warn workers around the B2200 about dangerous conditions.

Installation

Page 37 of the Agilent Technologies B2201A user manual — Power Cable
Page 37 · click to enlarge
Page 37

Requirements

This section describes the following requirements for the Agilent B2200.

  • “Power Requirements”
  • “Power Cable”
  • “Operating Environment”
  • “Storage and Shipping Environment”

Power Requirements

The Agilent B2200 can operate from any single-phase ac power source supplying 90 to 264 V in the frequency range from 47 to 63 Hz. The maximum power consumption is 2 A/200 VA.

Power Cable

In accordance with international safety standards, this instrument is equipped with a three-wire power cable. When connected to an appropriate ac power outlet, this cable grounds the instrument frame. The type of power cable shipped with each instrument depends on the country of destination. See the following table for the part numbers of the power cables available.

If the plug on the cable does not fit the power outlet, or the cable is to be attached to a terminal block, cut the cable at the plug end and re-wire it. This work should be performed by a qualified electrician all local electrical codes being strictly observed.

The color coding used in the cable will depend on the cable supplied. If a new plug is to be connected, it must meet local safety requirements and include the following features:

  • Adequate load-carrying capacity.
  • Ground connection.
  • Cable clamp.

Installation Requirements

Page 38 of the Agilent Technologies B2201A user manual — Requirements
Page 38 · click to enlarge
Page 38
  • Plug: CEE 7 sheet VII, 250 V, 10 A
  • Cable: 8120-4519, 8120-1689
  • Plug: NEMA 5-15P, 125 V, 10 A
  • Cable: 8120-6825, 8120-1378, 8120-1395
  • Plug: AS 3112, 250 V, 10 A
  • Cable: 8120-4419, 8120-1369
  • Plug: BS 1363/A, 250 V, 10 A
  • Cable: 8120-4420, 8120-1351
  • Plug: NEMA 6-15P, 250 V, 10 A
  • Cable: 8120-3996, 8120-0698
  • Plug: IS 1293 and IS 6538, 250 V, 10 A
  • Cable: 8121-1582, 8120-4211
  • Plug: SEV 1011, 250 V, 10 A
  • Cable: 8120-2104
  • Plug: SR 107-2-D1, 250 V, 10 A
  • Cable: 8120-2956
  • Plug: JIS C 8303, 125 V, 12 A
  • Cable: 8121-0743, 8120-4753
  • Plug: IRAM 207310A, 250 V, 10 A
  • Cable: 8120-6870
  • Plug: CEI 23-16, 250 V, 10 A
  • Cable: 8120-6978
  • Plug: Israel SI 32, 250 V, 10 A
  • Cable: 8120-5182
  • Plug: CNS 10917-2, 125 V, 10 A
  • Cable: 8121-1635, 8120-6825
  • Plug: CS 0017:2003, 250 V, 10 A
  • Cable: 8120-8871, 8120-0674, 8121-1638
  • Plug: GB 1002 figure 3 , 250 V, 10 A
  • Cable: 8120-8376
  • Plug: SANS 164-1, 250 V, 10 A
  • Cable: 8121-0564, 8120-4211

Installation Requirements

Page 39 of the Agilent Technologies B2201A user manual — Storage and Shipping Environment
Page 39 · click to enlarge
Page 39

Operating Environment

The Agilent B2200 must be operated within the following environmental conditions:

  • Temperature: 5 ° C to 35 ° C
  • Humidity: 5 % to 70 % R.H., non-condensing
  • Altitude: 0 m to 2,000 m

Storage and Shipping Environment

The Agilent B2200 should be stored or shipped in environments within the following limits:

  • Temperature: - 20 ° C to 70 ° C
  • Humidity:

B2200A/B2210A: < 80 % R.H. (at 35 °C), < 60 % R.H. (at 65 °C), non-condensing

B2201A/B2211A< 80 % R.H. (at 65 ° C), non-condensing
  • Altitude: 0 m to 15,240 m

Protect the Agilent B2200 from temperature extremes to prevent condensation from forming inside the Agilent B2200. If condensation occurs, the Agilent B2200 may damage or may not satisfy the specifications. If the Agilent B2200 was placed at the high temperature and high humidity environment long time, the Agilent B2200 may degrade its performance.

The Agilent B2200 must be packed in the certain packing materials for protection from damage when it is shipped. Cover the Agilent B2200 by using the moisture-proof and dehumidifying packing material (furnished), and pack it by using packing materials such as carton box and cushioning materials.

Installation Requirements

Page 40 of the Agilent Technologies B2201A user manual — Inspection
Page 40 · click to enlarge
Page 40

Inspection

The Agilent B2200 contains the condensation sensitive electronic parts. The condensation will have a negative impact on the Agilent B2200 to operate normally.

Do not open the packing materials, and leave the Agilent B2200 to acclimate it to the installation environment (temperature and humidity). If it is opened without enough acclimation, the Agilent B2200 may damage.

When you open the box that contains the Agilent B2200, check the following:

1. Before unpacking any components, inspect all boxes for any signs of damage that might have occurred during shipment such as:
  • Dents
  • Scratches
  • Cuts
  • Water marks
2. When you open the boxes that contain the Agilent B2200, check the components against the contents lists that are attached to the boxes.

If anything is wrong, notify your local Agilent Technologies sales office.

Installation Inspection

Page 41 of the Agilent Technologies B2201A user manual — Installing the B2200
Page 41 · click to enlarge
Page 41

Installation Installing the B2200

Installing the B2200

This section describes how to install the Agilent B2200.

1. See “Requirements” on page 2-3, and determine the installation location.
2. Unpack the Agilent B2200 and place it at the installation site.
3. Plug the power cable into the Line input receptacle at the rear panel.
4. Plug the power cable into the power receptacle.
5. Perform the self-test. See “Self-Test” on page 2-10.
6. If you control the Agilent B2200 by using an external computer:

Set the GPIB address. See “To Set the GPIB Address” on page 2-8 and “To Connect the GPIB Cable” on page 2-8.

7. If you have to add or replace the switch module:

Turn the Agilent B2200 off, remove the power cable, and install the switch module. See “To Install the Switch Module” on page 2-9 and “To Install the Blank Panel” on page 2-9.

If you require that the modules be guaranteed to their specifications (for ISO compliance, etc.), then you must perform a calibration on the instrument (modules and mainframe together). For calibration, contact your nearest Agilent Technologies service center.

Page 42 of the Agilent Technologies B2201A user manual — To Connect the GPIB Cable
Page 42 · click to enlarge
Page 42

Installation Installing the B2200

To Set the GPIB Address

Every device on the GPIB bus must have a unique address. If you need to change the GPIB address, turn the Agilent B2200 on and perform the following procedure. The new GPIB address is recognized only at power on. The Agilent B2200 leaves the factory with the GPIB address set to 22.

1. Press the Menu key.
2. Move the cursor to CONFIG, then press the Enter key.
3. Move the cursor to ADDRESS, then press the Enter key. The following message will appear on the LCD. XX will be 0 to 30.

GPIB Address = XX

4. Press the arrow key (up or down) to specify the desired GPIB address, then press the Enter key.
5. Press the Enter key twice.
6. Turn the Agilent B2200 off, and then turn it on again.

To Connect the GPIB Cable

To connect the instrument with a computer or peripheral device via GPIB (IEEE Std. 488), connect an GPIB cable between the GPIB connector on the instrument rear panel and the GPIB connector on the peripheral device.

Including the controller, a total of 15 GPIB interfaces can be connected on the same GPIB bus. The following are rules for connecting GPIB interfaces:

  • If total number of interfaces ≤ 10:

Max. total cable length = number of interfaces × 2 m.

  • If total number of interfaces > 10:

Max. total cable length = 20 m.

  • Maximum cable length between interfaces is 4 m.
  • Star connection and cascade connection are allowed. Loop connection is not allowed.
Page 43 of the Agilent Technologies B2201A user manual — To Install the Blank Panel
Page 43 · click to enlarge
Page 43

Installation Installing the B2200

To Install the Switch Module

Use clean handling and anti-static procedures when removing, configuring, and installing the switch modules. The modules contain components that can be damaged by static electricity.

The following procedure explains the module installation and removal.

1. Turn off the Agilent B2200, then wait at least 10 seconds before you remove or install a module.
2. Loosen the screws on both the left and right edges of a blank panel or a module attached to the slot you want to install a new module.
3. Remove the blank panel or the module.
4. Align the new module with the left and right slot guide rails. Then the component side should be facing up.
5. Push the module into the slot until you feel it seat firmly into the connector at the back of the slot.
6. Screw in the screws on the left and right edges of the module.
7. Execute the self-test and the relay test. See “Self-Test” on page 2-10.

To Install the Blank Panel

If the blank panel is not installed to cover an unused slot, install the blank panel as follows:

1. Align the blank panel over the unused slot.
2. Screw in the screws on the left and right edges of the blank panel.
Page 44 of the Agilent Technologies B2201A user manual — Self-Test
Page 44 · click to enlarge
Page 44

Self-Test

The self-test and diagnostics checks the operation of the mainframe and the modules. However they cannot confirm if the Agilent B2200 satisfies its specifications.

For verifying the specifications, contact your nearest Agilent Technologies Service Center. Trained service personnel will perform calibration (performance verification).

It is recommended to perform calibration once a year at least.

The following procedure performs the self-test and diagnostics.

1. Turn on the Agilent B2200.

The controller test will be performed. If the Agilent B2200 fails the test, contact your nearest Agilent Technologies Service Center.

2. Press the Shift key and the Menu key in this order. Self-test menu will appear on the LCD.

KEY Front panel interface test

LED LED matrix test

PEN Light pen test

3. Move the cursor to the test item to perform, then press Enter key.
4. To start the test, move the cursor to EXECUTE, then press Enter key.

To display the previous test result, move the cursor to RESULT, then press Enter key.

For the details of each test, see “Selftest Menu” on page 3-34.

Show 5 more labels / short text on this page
Installation Self-Test · SELF_TEST Controller test · RELAY_TEST Relay test · BEEPER Beeper test · GPIB GPIB test
Page 45 of the Agilent Technologies B2201A user manual — GUARD
Page 45 · click to enlarge
Page 45

Output Connections

This section describes how to connect the Agilent B2200 outputs to prober, connector plate, test fixture, and so on (DUT interface).

  • “Output Connectors”
  • “Connector Plates”
  • “To Make Connections to DUT Interface”
  • “To Make Interlock Circuit”
  • “To Mount Connectors”

If you do not use the connector plate for the connection between the output and the DUT interface, see “To Mount Connectors” on page 2-18.

Output Connectors

The Agilent B2200 output connectors are the triaxial BNC connector. The input signals appears at the output as shown in Figure 2-1 when input-output switching path is made. Figure 2-1 shows signals appear on the output connectors with non-Kelvin connection. If you make Kelvin connection, the signal of the even output connectors is Sense, not Force.

Figure 2-1 Output Connector and Output Signal

GUARD

GND

GND

(a) When SMU is connected (b) When AUX is connected

Installation Output Connections · Signal (force) · Signal (force)

Page 46 of the Agilent Technologies B2201A user manual — Connector Plates
Page 46 · click to enlarge
Page 46

Connector Plates

Connector plates (Table 2-2) are used for the connection between the Agilent B2200 outputs and the DUT interface (prober and so on). To connect to the connector plate, use the cable shown in Table 2-3.

Table 2-2 Connector Plate

Agilent Model No. Description

16495F Half size connector plate

16495F-001 has 12 triaxial through connectors (female to female), an Intlk connector, and a GNDU connector (triaxial through, female to female). The back of the Intlk connector is designed for soldering.

16495F-002 has 12 triaxial connectors, an Intlk connector, and a GNDU connector. The back of each connector is designed for soldering.

16495G Full size connector plate

16495G-001 has 24 triaxial through connectors (female to female), an Intlk connector, and a GNDU connector (triaxial through, female to female). The back of the Intlk connector is designed for soldering.

16495G-002 has 24 triaxial connectors, an Intlk connector, and a GNDU connector. The back of each connector is designed for soldering.

This plate is used to cover openings when you made too big openings for mounting the connector plate. You will use this plate to cover openings if you mount the half size connector plate in openings made for the full size connector plate.

Table 2-3 Output Cable

Agilent Model No. Description

16494A Triaxial cable (for non-Kelvin connection)

16494B Kelvin Triaxial cable (for Kelvin connection)

To install the connector plate, refer to Agilent 16495 Installation Guide.

Installation Output Connections · 16495E Blank plate

Page 47 of the Agilent Technologies B2201A user manual — To Make Connections to DUT Interface
Page 47 · click to enlarge
Page 47

For Kelvin connection, use Kelvin triaxial cable listed in Table 2-3. To make a Kelvin output port (1, 3, 5, 7, 9 and 11), couple two E5252A output ports as follows:

Kelvin Output Port Output Port Number

1 1 (Force) and 2 (Sense)

3 3 (Force) and 4 (Sense)

5 5 (Force) and 6 (Sense)

7 7 (Force) and 8 (Sense)

9 9 (Force) and 10 (Sense)

11 11 (Force) and 12 (Sense)

To Make Connections to DUT Interface

This section describes for the connections between the DUT interface and the connectors connected to the Agilent B2200 output cables. See Table 2-4.

For the extended measurement paths over the connector plate, use low-noise coaxial cable (Agilent part number 8121-1191). This cable can maximize the guard effects and minimize the impression of the external noise.

Figure 2-2 shows the cutting example of this cable. Key point is the isolation between the conductive layer and the center conductor. So, cut and trim the end of the cable as shown in this figure by using a cutter and so on.

Figure 2-2 Coaxial Cable Cutting Example

Cover here using sleeve

Center conductor (for Force/Sense signal)

approx. 10 to 15 mm

Outer conductor (for Guard signal) Insulator (clear)

Show 4 more labels / short text on this page
Installation Output Connections · Insulator (black) · min 2 mm · Conductive layer (black)
Page 48 of the Agilent Technologies B2201A user manual — To Make Connections to DUT Interface
Page 48 · click to enlarge
Page 48

Table 2-4 Connection to the DUT Interface

Agilent Technologies B2201A — To Make Connections to DUT Interface diagram (page 48)
Fig. 6 · To Make Connections to DUT Interface

This connection is available only for the Kelvin connectors.

Following connection is for the Kelvin connectors. For the triaxial connectors, ignore SENSE terminal, and make connection only for FORCE terminal. Measurement data will include residual resistance from the connection cable.

This connection can cancel effects of cable resistance by connect sense line and force line as close as possible to DUT terminal.

For highly accurate current forcing and measurements while minimizing leakage, surround all force and sense lines from SMU by a guard as far as possible, and make cables stable by taping.

For reducing capacitance measurement error, tape the cables to any grounded materials such as shielding box. If you use probe card, grounding the probe card also reduces capacitance measurement error.

Installation Output Connections

Page 49 of the Agilent Technologies B2201A user manual — To Install Interlock Circuit
Page 49 · click to enlarge
Page 49

To Make Interlock Circuit

The interlock circuit is to prevent electric shock when touching measurement terminals. You must install an interlock circuit on shielding box to prevent dangerous voltages when door of the shielding box is open.

Figure 2-3 shows the pin assignments of the interlock connector.

Figure 2-3 Interlock Connector Pin Assignments

To Install Interlock Circuit

Install the interlock circuit as follows.

1. Mount two mechanical switches on your shielding box, so that the switches close when the door of the shielding box is closed, and open when the door is opened. For the recommended parts and the dimensions of the switch, see Figure 2-4 and Figure 2-5.
2. Use wire to connect the two switches in series between pin number 1 and 2 (or 3) of the interlock connector. See Figure 2-3.

For example, Agilent 4155/4156 is connected to the interlock circuit, it cannot force more than ±40 V when the door is open. When door is closed, it can force more than ±40 V.

Installation Output Connections

Page 50 of the Agilent Technologies B2201A user manual — To Install LED Circuit
Page 50 · click to enlarge
Page 50
Agilent Technologies B2201A — To Make Interlock Circuit diagram (page 50)
Fig. 7 · To Make Interlock Circuit

Figure 2-4 Dimensions of Interlock Switch (Agilent part number 3101-0302)

Figure 2-5 Dimensions of Interlock Switch (Agilent part number 3101-3241)

Agilent Technologies B2201A — To Make Interlock Circuit diagram (page 50)
Fig. 8 · To Make Interlock Circuit

To Install LED Circuit

To install LED circuit on your shielding box, do following:

1. Mount LED on your shielding box. For the recommended parts and the dimensions of the LED, see Figure 2-6.
2. Use wire to connect the LED between pin 4 and pin 5 (or 6) of the interlock connector. See Figure 2-3.
Page 51 of the Agilent Technologies B2201A user manual — To Connect Interlock Circuit to Instrument
Page 51 · click to enlarge
Page 51

The 4155/4156 semiconductor parameter analyzer's Intlk connector provides the interlock signal and a LED drive signal. If a LED is connected between pin 4 and pin 5 (or 6) of the interlock connector, the LED lights to indicate high voltage output when more than ±40 V is forced from an SMU in the 4155/4156.

Figure 2-6 Dimensions of LED (Agilent part number 1450-0641)

Unitsmm

To Connect Interlock Circuit to Instrument

Before beginning the measurement, connect the interlock circuit to the instrument’s interlock connector as follows.

  • For the instruments which has a BNC-Type interlock connector:
1. Get the following parts.
  • Agilent 16493J Interlock cable, 1 ea.
  • Agilent 16435A Interlock cable adapter, 1 ea.
2. Connect the 16493J interlock cable between the interlock circuit and the 16435A adapter.
3. Connect the BNC cable (furnished with the adapter) between the adapter and the instrument’s interlock connector.
  • For Agilent 4155/4156/E5260/E5270:

Connect the 16493J interlock cable between the interlock circuit and the instrument’s interlock connector. You can connect it directly without using any adapter.

Installation Output Connections · Cathode(-) · Anode(+)

Page 52 of the Agilent Technologies B2201A user manual — To Mount Connectors
Page 52 · click to enlarge
Page 52

If you use the 4155/4156, you can easily check the interlock circuit as follows:

1. Connect the Intlk connector of the 4155/4156 to your interlock circuit.
2. Press System front-panel key, then select CALIB/DIAG primary softkey to display the SYSTEM: SELF-CALIBRATION/DIAGNOSTICS page.
3. In the CALIB/DIAG field, select DIAG secondary softkey.
4. In the CATEGORY field, select I/O PERIPH secondary softkey.
5. Move pointer to the 403 (INT.) Interlock LED field.
6. Select EXECUTE secondary softkey.
7. Confirm the following:
  • LED turns on within 1 sec from when interlock circuit is shorted.
  • LED turns off within 1 sec from when interlock circuit is open.

To stop the interlock test, select STOP secondary softkey.

To Mount Connectors

This section provides the information needed to mount connectors for the Agilent B2200 output cable connections on the shielding panel. See this section if you do not use the connector plate.

1. Get the appropriate parts for your situation. Refer to Table 2-5.
2. Make holes and mount the connectors. Refer to Table 2-6.

For kelvin connections, use the Kelvin cables. The Kelvin cable requires the kelvin triaxial connector which has two connector holes and three screw holes.

3. Make an interlock circuit as shown in “To Make Interlock Circuit” on page 2-15.
4. Make cable connections to the DUT interface. Refer to “To Make Connections to DUT Interface” on page 2-13.

Installation Output Connections

Page 53 of the Agilent Technologies B2201A user manual — To Mount Connectors
Page 53 · click to enlarge
Page 53

Table 2-5 Recommended Parts

Description Agilent Part No.

Interlock Connector (6 pin, female) 1252-1419C

Switch 3101-0302 or 3101-3241

LED (VF @ 2.1 V @ IF = 10 mA) 1450-0641

Triaxial Connector (female) 1250-2457

Low Noise Coaxial Cable 8121-1191

Table 2-6 Dimensions of Connector Holes

Triaxial Connector (in mm) Interlock Connector (in mm)

∅11.3

Kelvin Triaxial Connector (in mm)

14 11 11 14 2.8

2 − ∅11.3 ∅3.2 2 − M3 x 0.5

Installation Output Connections · Wire 8150-5680

Page 54 of the Agilent Technologies B2201A user manual — Input Connections
Page 54 · click to enlarge
Page 54

Input Connections

This section explains how to connect instruments to the Agilent B2200.

The Agilent B2200 input connector has eight SMU input connectors and six AUX input connectors. Table 2-7 shows cables used to connect the instrument for typical applications using the Agilent B2200.

You can make the Kelvin connections when using the E5260/E5270/4142/4156 SMU or 41501 HPSMU. Then use cables as follows:

  • Agilent 16493K Kelvin triaxial cable for Agilent E5260/E5270/4156/41501
  • Agilent 16494B Kelvin triaxial cable for Agilent 4142B

To connect the Kelvin triaxial cable to the Agilent B2200 input, use pairs (couple ports) of the SMU input connectors (1&2, 3&4, 5&6, or 7&8). The input signal for each Kelvin input is as follows:

  • Odd input port number: Force
  • Even input port number: Sense

Ground Unit (GNDU) can sink over 1 A. For the measurement over 1 A, connect the GNDU to the connector plate directly by using the Agilent 16493H GNDU cable.

If you never perform the measurement over 1 A, you can connect the GNDU to the Agilent B2200 inputs. Then use the Agilent 16493N GNDU cable. This cable requires the Kelvin connection.

Installation Input Connections

Page 55 of the Agilent Technologies B2201A user manual — Input Connections
Page 55 · click to enlarge
Page 55

Table 2-7 B2200 Input Connections

Cable Required Agilent Model No.

4155 MPSMU SMU1 to 8 16494A triaxial cable

VSU AUX1 to 6 BNC cable

VMU AUX1 to 6 BNC cable

Intlk − 16493J interlock cable (for direct connection to connector plate)

4156 HRSMU SMU1 to 8 16493K Kelvin triaxial cable or 16494A triaxial cable

VSU AUX1 to 6 BNC cable

VMU AUX1 to 6 BNC cable

Intlk − 16493J interlock cable (for direct connection to connector plate)

41501 HPSMU SMU1 to 8 16493K Kelvin triaxial cable or 16494A triaxial cable

MPSMU SMU1 to 8 16494A triaxial cable

PGU AUX1 to 6 BNC cable

GNDU SMU1 to 8 16493N GNDU Kelvin cable

− 16493H GNDU cable (for direct connection to connector plate)

4142B SMU SMU1 to 8 16494B Kelvin triaxial cable or 16494A triaxial cable

VSU AUX1 to 6 BNC cable

VMU AUX1 to 6 BNC cable

GNDU SMU1 to 8 16493HN GNDU Kelvin cable

− 16493H GNDU cable (for direct connection to connector plate)

Intlk − 16435A interlock cable adapter and 16493J interlock cable (for direct connection to Connector Plate)

Show 6 more labels / short text on this page
Installation Input Connections · Instrument B2200 · Application · Input Connector · Output Connector · DC Measurement
Page 56 of the Agilent Technologies B2201A user manual — E5260/ E5270
Page 56 · click to enlarge
Page 56

Cable Required Agilent Model No.

E5260/ E5270

SMU SMU1 to 8 16493K Kelvin triaxial cable or 16494A triaxial cable

GNDU SMU1 to 8 16493N GNDU Kelvin cable

− 16493H GNDU cable (for direct connection to connector plate)

Interlock − 16493J interlock cable (for direct connection to connector plate)

C Meter High, Low CMH/CML 16494F CMU cable for 4 terminal pairs

Pulse Input Pulse Generator

Output AUX1 to 6 BNC cable

Bias Input Power Supply

Output AUX1 to 6 BNC cable

In the table above,

  • CMH/CML connectors are a pair of the AUX inputs.
  • The Agilent B2200 does not have the interlock connector. Connect directly from instrument to the connector plate which has an interlock connector.
  • The Agilent 41422A/41423A (HCU/HVU) of the Agilent 4142B cannot be used with the Agilent B2200.
  • To connect instrument output that uses banana plug to the AUX connector, use the dual banana plug to BNC adapter (Agilent part number 1251-2277).
Show 7 more labels / short text on this page
Installation Input Connections · Instrument B2200 · Application · Input Connector · Output Connector · DC Measurement · C Measurement
Page 57 of the Agilent Technologies B2201A user manual — Measurement Cable Length
Page 57 · click to enlarge
Page 57

Installation Measurement Cable Length

Measurement Cable Length

This section describes how to calculate the total guard capacitance when using an SMU (source monitor unit). When using an SMU, the length of measurement cables is limited by the guard capacitance of the cables. The guard capacitance means the capacitance between the signal line (Force or Sense) and the Guard line.

Table 2-8 lists the guard capacitance for each element of the Agilent B2200’s measurement environment.

When using the 4155/4156/4142B/E5260/E5270, the maximum limit of the guard capacitance is approximately 900 pF. So, you add the following and total must be less than 900 pF:

  • Guard capacitance of cable from SMU to B2200 inputs.
  • Guard capacitance of mainframe and modules.
  • Guard capacitance of cable from B2200 outputs.
  • Guard capacitance of cable from connector plate to DUT.
  • Other capacitances, such as for probe card.

The following is an example to calculate the guard capacitance of a measurement environment:

Installed Plug-in cardsFour B2210As (145 pF + 8 pF × 3)
Input Cable16494A-003 (75 pF)
Output Cable16494A-001 (130 pF)

Cable from Connector Plate to DUT: 8121-1191, 1 m (130 pF)

Probe CardGuard Capacitance approximately 10 pF (example)

In this environment, the total guard capacitance (Cg-total) is as follows.

Cg-total = 145+(8×3)+75+130+130+10 pF = 514 pF < 900 pF

Even if you use other instruments, you can use the table for reference about the measurement cable length and capacitance.

Page 58 of the Agilent Technologies B2201A user manual — Measurement Cable Length
Page 58 · click to enlarge
Page 58

Installation Measurement Cable Length

Table 2-8 Guard Capacitances of B2200 Measurement Environment (Typical)

Connection Agilent Model/Part No. Cable Length

from SMU to B2200 inputs

16494A-003 (for non-Kelvin) 80 cm 75 pF

16494A-001 (for non-Kelvin) 1.5 m 130 pF

16494B-003 (for Kelvin, 4142B) 80 cm 90 pF

16493K-001 (for Kelvin) 1.5 m 150 pF

B2210A (only 1 card is installed) − 145 pF

B2210A (more 1 card) − 8 pF/extra card

B2211A (only 1 card is installed) − 145 pF

B2211A (more 1 card) − 8 pF/extra card

from B2200 outputs to connector plate or probe card interface

16494A-001 (for non-Kelvin) 1.5 m 130 pF

16494A-002 (for non-Kelvin) 3 m 240 pF

16494A-003 (for non-Kelvin) 4 m 340 pF

16494B/C-001 (for Kelvin) 1.5 m 140 pF

16494B/C-002 (for Kelvin) 3 m 260 pF

16494C-003 (for Kelvin) 4 m 340 pF

from connector plate to DUT

Agilent Part No. 8121-1191 X m 130 pF per m

probe card interface B2220A − 70 pF

Guard Capacitance · mainframe and modules

Page 59 of the Agilent Technologies B2201A user manual — Cleaning
Page 59 · click to enlarge
Page 59

Maintenance

Maintenance should be performed periodically to keep the B2200 in good condition.

Calibration

Calibration must be performed periodically so that the instruments satisfy the specifications, and keep a good condition. It is recommended to perform a calibration once a year at least. For calibration, contact your nearest Agilent Technologies Service Center. Trained service personnel will perform calibration (performance verification).

Cleaning

Before performing cleaning, turn off the instrument, and disconnect power cable from the rear panel. Use a dry cloth to clean the external case parts.

To prevent electrical shock, do not perform cleaning when the instrument is turned on, and do not use a wet cloth.

Installation Maintenance

Page 60 of the Agilent Technologies B2201A user manual — Maintenance
Page 60 · click to enlarge
Page 60

Installation Maintenance

Page 61 of the Agilent Technologies B2201A user manual — 3 Front Panel Operation
Page 61 · click to enlarge
Page 61

3 Front Panel Operation

Page 62 of the Agilent Technologies B2201A user manual — 3 Front Panel Operation
Page 62 · click to enlarge
Page 62

This chapter explains the front panel operation and the switch control functions of the Agilent B2200, also provides the reference information of the front panel keys and display.

  • “Operation”
  • “Switch Control Functions”
  • “Display Functions”
  • “Front Panel Keys”
  • “Setup Menus”

Front Panel Operation

Page 63 of the Agilent Technologies B2201A user manual — Operation
Page 63 · click to enlarge
Page 63

Operation

This section describes operations of the Agilent B2200.

  • “To Initialize Agilent B2200”
  • “To Enable Light Pen”
  • “To Change Channel Configuration Mode”
  • “To Change Connection Rule”
  • “To Change Connection Sequence”
  • “To Control Switch Condition”
  • “To Open All Switches”
  • “To Save/Load Setup Data”
  • “To Use Bias Mode”
  • “To Use Ground Mode”
  • “To Use Couple Mode”
  • “To Display Firmware Revision”
  • “To Display Module Information”
  • “To Read Error Message”
  • “To Set Beeper”
  • “To Set GPIB Address”
  • “To Set Remote Display Mode”
  • “To Return to Local Mode”

Use the right/left arrow key to move the cursor to the desired function name. Use the up/down arrow key to select the value for the setup item.

The Enter key makes the setup effective, and returns the display to the previous menu or display.

The Exit key returns the display to the previous menu or display without making the setup effective.

Front Panel Operation · Operation

Page 64 of the Agilent Technologies B2201A user manual — To Change Channel Configuration Mode
Page 64 · click to enlarge
Page 64

Front Panel Operation Operation

To Initialize Agilent B2200

1. Press the Shift key and the Local key. Then the LCD will display NO.
2. Press the arrow key to displays YES, then press the Enter key to initialize the Agilent B2200, or press the Exit key to cancel initialization.

To Enable Light Pen

Turn the Agilent B2200 off, and connect the light pen to the Light Pen connector at the right down corner of the front panel. After that, perform the following procedure to enable the light pen.

1. Press the Menu key. The setup menu will be displayed on the LCD.
2. Press the arrow key to move the cursor to PEN, then press the Enter key.
3. Press the arrow key to display ENABLE, then press the Enter key to enable the light pen, or press the Exit key to cancel changing the setup.

You can control the Agilent B2200 functions and the switch connections by using the front panel keys. And the Agilent B2200 allows you to use the Agilent 16443A Light Pen to control the switch connections.

Pointing a LED on the LED matrix by using the light pen controls the setting of the switch specified by the LED.

To Change Channel Configuration Mode

1. Press the Shift key and the Bias Mode key. The LCD displays the present setting (AUTO or NORMAL).
2. Press the arrow key to select the mode, then press the Enter key. The LCD displays NO.
3. Press the arrow key to display YES, then press the Enter key. Or press the Exit key to cancel this operation.

Changing the channel configuration mode resets the Agilent B2200.

Page 65 of the Agilent Technologies B2201A user manual — To Control Switch Condition
Page 65 · click to enlarge
Page 65

To Change Connection Rule

1. Press the Rule key. The LCD displays the present setting (FREE or SINGLE).
2. Press the arrow key to select the connection rule, then press the Enter key. Or press the Exit key to cancel this operation.

To Change Connection Sequence

1. Press the Shift key and the Rule key. The LCD displays the present setting (Break_Before_Make, Make_Before_Break, or No_Sequence).
2. Press the arrow key to select the connection sequence, then press the Enter key. Or press the Exit key to cancel this operation.

To Control Switch Condition

If you use the light pen, you do not need the following instruction to control the switch condition.

1. If you are NOT watching the orange-blink-LED on the LED matrix, press the

Open/Close key. The blink-LED will appear.

2. Press the arrow keys to move the blink-point.
3. Press the Open/Close key to open/close the switch specified by the blink-LED. The green-LEDs indicate the switches that are closed now.
4. Repeat 2 and 3 until the switch setup is completed.

Blinking of LED will be stopped by pressing the front panel key other than the

Open/Close, Fast, or arrow keys.

For the Kelvin outputs, the input-output paths must be controlled as shown below.

  • For the input couple port:

Control the switches to connect the input couple port to the Kelvin output.

  • For the single input port:

Control the switches to connect the input port to two ports of the Kelvin output.

If you do not comply with this note, leakage current between the Kelvin paths will cause measurement error.

Front Panel Operation · Operation

Page 66 of the Agilent Technologies B2201A user manual — To load the setup
Page 66 · click to enlarge
Page 66

Front Panel Operation Operation

To Open All Switches

1. Press the Open All key.
2. Press the arrow key to display YES, then press the Enter key.

When the bias mode is ON, the bias-enabled output ports will be connected to the input bias port.

When the ground mode is ON, the ground-enabled input/output ports will be connected to the input ground port.

To Save/Load Setup Data

The Agilent B2200 provides the internal memory used to save the setup condition. Eight setup conditions can be saved into the memory.

To save the setup

1. Press the Shift key and the Load Memory key.
2. Press the arrow key to select the internal memory (01 to 08).
3. Press the Enter key to save the setup. Or press the Exit key to cancel this operation.

Wait until the save operation is completed. Then, the LCD displays the status and setup conditions.

To load the setup

1. Press the Load Memory key.
2. Press the arrow key to select the internal memory (01 to 08).
3. Press the Enter key to load the setup. Or press the Exit key to cancel this operation. Wait until the load operation is completed. Then, the LCD displays the status and setup conditions.
Page 67 of the Agilent Technologies B2201A user manual — To Use Bias Mode
Page 67 · click to enlarge
Page 67

To Use Bias Mode

Bias mode cannot be set to ON when the ground mode is ON.

1. Change the bias-enabled output ports.

a. Press the Shift key and the Port Function key.

b. Move the cursor to SET, then press the Enter key.

c. Move the cursor to DISABLE, then press the Enter key.

The LCD displays one of the bias-enabled output port numbers which can be changed to the bias-disabled. Or if there is no bias-enabled output port, the LCD displays Channel: No Channel. Press the arrow key to select the output port to be changed to the bias-disabled, then press the Enter key.

d. Move the cursor to ENABLE, then press the Enter key.

The LCD displays one of the bias-disabled output port numbers which can be changed to the bias-enabled. Or if there is no bias-disabled output ports, the LCD displays Channel: No Channel. Press the arrow key to select the output port to be changed to the bias-enabled, then press the Enter key.

e. Repeat c and d until the bias-enabled port setup is completed. After that, press the Exit key more than once to display the status and setup conditions.

2. Press the Bias Mode key to set the bias mode ON. Pressing the key again sets the mode OFF.
3. Change the bias port (default setting: 10).

a. When the bias mode is ON, press the Port Function key. The blink-cursor appears on the first line of the LCD.

b. Press the right/left arrow key to move the cursor on the input port number

(01 to 14) that will be set to the input bias port.

c. Press the up/down key to display B, then press the Enter key.

Front Panel Operation · Operation

Page 68 of the Agilent Technologies B2201A user manual — To Use Ground Mode
Page 68 · click to enlarge
Page 68

Front Panel Operation Operation

To Use Ground Mode

Ground mode cannot be set to ON when the bias mode is ON.

1. Change the ground-enabled output ports.

a. Press the Shift key and the Open All key.

b. Move the cursor to SET, then press the Enter key.

c. Move the cursor to ENABLE, then press the Enter key.

The LCD displays one of the ground-disabled output port numbers which can be changed to the ground-enabled. Or if there is no ground-disabled output ports, the LCD displays Channel: No Channel. Press the arrow key to select the output port to be changed to the ground-enabled, then press the Enter key.

d. Move the cursor to DISABLE, then press the Enter key.

The LCD displays one of the ground-enabled output port numbers which can be changed to the ground-disabled. Or if there is no ground-enabled output port, the LCD displays Channel: No Channel. Press the arrow key to select the port to be changed to the ground-disabled, then press the Enter key.

e. Repeat c and d until the ground-enabled port setup is completed. After that, press the Exit key more than once to display the status and setup conditions.

2. Press the Shift key and the Couple Mode key to set the ground mode ON. Performing this operation again sets the mode OFF.
3. Change the ground port (default setting: 12).

a. When the ground mode is ON, press the Port Function key. The blink-cursor appears on the first line of the LCD.

b. Press the right/left arrow key to move the cursor on the input port number

(01 to 14) that will be set to the input ground port.

c. Press the up/down key to display G.

4. Change the ground-enabled input ports (multiple ports can be set). Repeat a and b for the all ground-enabled input ports, then press the Enter key.

a. Press the right/left arrow key to move the cursor on the input port number (01 to 08) that will be set to the ground-enabled input port.

b. Press the up/down key to display -.

Open the ground-enabled input ports to prevent the instrument from damage.

Page 69 of the Agilent Technologies B2201A user manual — To Display Module Information
Page 69 · click to enlarge
Page 69

To Use Couple Mode

1. Press the Couple Mode key to set the couple mode ON. Pressing the key again sets the mode OFF.
2. Perform the input couple port detection.

a. Press the Menu key. The setup menu will be displayed on the LCD.

b. Move the cursor to SCAN, then press the Enter key. The LCD displays NO.

c. Press the arrow key to display YES, then press the Enter key.

d. Press the Exit key.

3. Change the couple ports (multiple ports can be set).

a. When the couple mode is ON, press the Port Function key. The blink-cursor appears on the first line of the LCD.

b. Press the right/left arrow key to move the cursor on the input port number

(01 to 14) that will be set to the input couple port.

c. Press the up/down key to display C.

d. Repeat b and c for the all couple ports, then press the Enter key.

To Display Firmware Revision

1. Press the Menu key. The setup menu will be displayed on the LCD.
2. Move the cursor to CONFIG, then press the Enter key.
3. Move the cursor to REVISION, then press the Enter key. The firmware revision is displayed.

To Display Module Information

1. Press the Menu key. The setup menu will be displayed on the LCD.
2. Move the cursor to CONFIG, then press the Enter key.
3. Move the cursor to UNIT, then press the Enter key. The module information is displayed.
4. Press the arrow keys to display another module information item.

Front Panel Operation · Operation

Page 70 of the Agilent Technologies B2201A user manual — To Set GPIB Address
Page 70 · click to enlarge
Page 70

Front Panel Operation Operation

To Read Error Message

1. Press the Menu key. The setup menu will be displayed on the LCD.
2. Move the cursor to ERROR, then press the Enter key.
3. Move the cursor to DISPLAY, then press the Enter key to display the message.
4. Press the arrow keys to read another error message. A maximum of four error messages can be stored.

To clear error buffer

1. Press the Menu key. The setup menu will be displayed on the LCD.
2. Move the cursor to ERROR, then press the Enter key.
3. Move the cursor to CLEAR, then press the Enter key. The LCD displays NO.
4. Press the arrow key to display YES, then press the Enter key to clear the error buffer, or press the Exit key to cancel operation.

To Set Beeper

1. Press the Menu key. The setup menu will be displayed on the LCD.
2. Move the cursor to BEEP, then press the Enter key.
3. Press the arrow keys to select the beep ON or OFF.
4. Press the Enter key to make the setup effective, or press the Exit key to cancel changing the setup.

To Set GPIB Address

1. Press the Menu key. The setup menu will be displayed on the LCD.
2. Move the cursor to CONFIG, then press the Enter key.
3. Move the cursor to ADDRESS, then press the Enter key.
4. Press the arrow keys to set the GPIB address.
5. Press the Enter key to make the setup effective, or press the Exit key to cancel changing the setup. To be effective the new address, reboot the Agilent B2200.
Page 71 of the Agilent Technologies B2201A user manual — To Return to Local Mode
Page 71 · click to enlarge
Page 71

To Set Remote Display Mode

This instruction enables or disables the data display in the GPIB remote condition. See “RMT_DSPL” on page 3-33.

LCD

1. Press the Menu key. The setup menu will be displayed on the LCD.
2. Move the cursor to RMT_DSPL, then press the Enter key.
3. Move the cursor to LCD, then press the Enter key.
4. Press the arrow keys to set the remote display mode ON or OFF.
5. Press the Enter key to make the setup effective, or press the Exit key to cancel changing the setup.

LED matrix

1. Press the Menu key. The setup menu will be displayed on the LCD.
2. Move the cursor to RMT_DSPL, then press the Enter key.
3. Move the cursor to LED, then press the Enter key.
4. Press the arrow keys to set the remote display mode ON or OFF.
5. Press the Enter key to make the setup effective, or press the Exit key to cancel changing the setup.

To Return to Local Mode

Press the Local key. If the front panel keys are locked, send the :SYST:KLC command from an external computer, then press the Local key.

Front Panel Operation · Operation

Page 72 of the Agilent Technologies B2201A user manual — Switch Control Functions
Page 72 · click to enlarge
Page 72

Front Panel Operation Switch Control Functions

Switch Control Functions

This section introduces the switch control functions of the Agilent B2200.

  • “Channel Configuration Mode”
  • “Connection Rule”
  • “Connection Sequence”
  • “Bias Mode”
  • “Ground Mode”
  • “Couple Mode”
Page 73 of the Agilent Technologies B2201A user manual — Channel Configuration Mode
Page 73 · click to enlarge
Page 73

Front Panel Operation Switch Control Functions

Channel Configuration Mode

The Agilent B2200 provides the two channel configuration modes, Normal and Auto. The configuration mode defines the way to control multiple switch modules installed in the mainframe.

Normal Each module is considered independently. So each module is always a 12 output switching matrix.

The switch control functions need to be defined for each module.

Auto Default mode at Power On or *RST command.

To use this mode, the same type of module must be installed in the mainframe from slot number 1 continuously (slots 1 to 2; or slots 1 to 3; or slots 1 to 4).

The installed modules are treated as one switching matrix. For example, if the modules are installed in slots 1 to 3, the Agilent B2200 works as a 36 output switching matrix.

The switch control functions are available for the switching matrix.

Page 74 of the Agilent Technologies B2201A user manual — Connection Rule
Page 74 · click to enlarge
Page 74

Front Panel Operation Switch Control Functions

Connection Rule

Connection rule is used to specify if an input or output port can have multiple connections. For each module installed in the mainframe (for auto configuration mode, installed modules are treated as one module), you can specify one of the following connection rules:

Single: Each input port can be connected to only one output port, and each output port can be connected to only one input port. So, existing connection to a port will be disconnected when a new connection is made.

If normal configuration mode is set to the Agilent B2200 with multiple modules, an input/output path is effective for each module. For example, if normal configuration mode is set to the Agilent B2200 with two modules, and single rule is set to both modules, the following connection is possible.

1 2 12 3 3

Free: Each input port can be connected to multiple output ports, and each output port can be connected to multiple input ports.

Show 4 more labels / short text on this page
Input · Close Open · Module 1 Output · Module 2 Output
Page 75 of the Agilent Technologies B2201A user manual — Connection Sequence
Page 75 · click to enlarge
Page 75

Front Panel Operation Switch Control Functions

Connection Sequence

Connection sequence specifies the open/close sequence of the relays when changing from an existing connection to a new connection.

You can select one of the three connection sequences shown in Table 3-1 for each module installed in the Agilent B2200.

When an existing connection is changed to a new connection, the relays are opened/closed as follows:

Table 3-1 Connection Sequence

Break Before Make 1. Disconnect previous route. 2. Wait for relays to open. 3. Connect new route.

Make Before Break 1. Connect new route. 2. Wait for relays to close. 3. Disconnect previous route.

No Sequence 1. Disconnect previous route. 2. Connect new route.

Connection · Sequence Relay Operation

Page 76 of the Agilent Technologies B2201A user manual — Bias Mode
Page 76 · click to enlarge
Page 76

Front Panel Operation Switch Control Functions

Bias Mode

Bias mode is useful for connecting the same input to multiple channels (output ports) simultaneously. And the connections will be kept until the other input port connection will be changed.

When the bias mode is ON, the input bias port is connected to all bias enabled output ports that are not connected to any other input ports. You cannot directly control which output ports are connected to the input bias port:

  • If another input port is disconnected from a bias enabled output port, the output port is automatically connected to the input bias port.
  • If another input port is connected to a bias enabled output port, the output port is automatically disconnected from the input bias port.
  • Bias disabled output ports are never connected to the input bias port when the bias mode is ON.
  • You can independently set the bias port and control the bias mode for each module when the configuration mode is normal.

When the bias mode is OFF, the input bias port is the same as the other input ports, so relays can be controlled directly to connect to output ports.

Figure 3-1 shows an example. Input 2 is the input bias port. When the bias mode is turned ON, all bias enabled output ports (that are not connected to other input ports) are connected to the input bias port. So, the output ports that are connected to input port 1 and 3 are not connected to the input bias port.

Figure 3-1 Example of Bias Mode

Agilent Technologies B2201A — Bias Mode diagram (page 76)
Fig. 9 · Bias Mode

Open Connects to bias port Connects to input 01

Page 77 of the Agilent Technologies B2201A user manual — Bias Mode
Page 77 · click to enlarge
Page 77

Front Panel Operation Switch Control Functions

Connection sequence (to connect input bias port to output ports) is always Break-Before-Make.

Bias mode cannot be set to ON when the ground mode is ON.

If the bias input port and a couple port have been assigned to the same input port, the bias mode and the couple mode cannot be used in parallel.

Page 78 of the Agilent Technologies B2201A user manual — Ground Mode
Page 78 · click to enlarge
Page 78

Front Panel Operation Switch Control Functions

Ground Mode

Ground mode operation is similar to the bias mode operation. The ground mode is useful for connecting the same input to multiple channels (output ports) and the unused input ports simultaneously. And the connections will be kept until the other input port connection will be changed. The ground mode is especially useful for settling the potential of the unused input/output paths.

If you assign the input port 12 (AUX Input 12) as the ground port, open the input port 12. This ground port will be internally connected to the ground when the ground mode is set to ON.

When the ground mode is ON, the input ground port is connected to all ground enabled input ports (unused ports) and the ground enabled output ports that are not connected to any other input ports. You cannot directly control which output ports are connected to the input ground port:

  • If another input port is disconnected from a ground enabled output port, the output port is automatically connected to the input ground port.
  • If another input port is connected to a ground enabled output port, the output port is automatically disconnected from the input ground port.
  • Ground disabled output ports are never connected to the input ground port when the ground mode is ON.
  • You can independently set the ground port and control the ground mode for each module when the configuration mode is normal.

When the ground mode is OFF, the input ground port is the same as the other input ports, so relays can be controlled directly to connect to output ports.

Connection sequence (to connect input ground port to output ports) is always Break-Before-Make.

Ground mode cannot be set to ON when the bias mode is ON.

The input ground port and a ground enabled input port cannot be assigned to the same input port.

If the ground input port and a couple port have been assigned to the same input port, the ground mode and the couple mode cannot be used in parallel.

If a ground enabled input port and a couple port have been assigned to the same input port, the ground mode and the couple mode cannot be used in parallel.

Page 79 of the Agilent Technologies B2201A user manual — Ground Mode
Page 79 · click to enlarge
Page 79

Front Panel Operation Switch Control Functions

Open the ground enabled input ports (unused input ports). If any equipment is connected to the ground enabled input ports, turning the ground mode ON may cause damage in the equipment.

Page 80 of the Agilent Technologies B2201A user manual — 5 SMU 5, 6
Page 80 · click to enlarge
Page 80

Front Panel Operation Switch Control Functions

Couple Mode

Couple mode is useful for making the Kelvin connections. When the couple mode is ON, the input couple ports will be connected to the output couple ports. For the available couple ports, see Table 3-2.

  • For the couple port connections, the odd/even number of the input port will be connected to the odd/even number of the output port respectively. Then the even number is always the odd number plus 1.
  • You can independently specify the same or different couple ports and control the couple mode for each module when the configuration mode is normal.

Table 3-2 Input Couple Port Numbers and Output Couple Port Numbers

Available input couple port No. Input ports Available output couple port No.

1 SMU 1, 2 1 and 2, 3 and 4, . . . , 11 and 12 (for normal configuration mode)

3 SMU 3, 4

1 and 2, 3 and 4, . . . , 47 and 48 (for auto configuration mode)

5 SMU 5, 6

7 SMU 7, 8

9 AUX 9, 10

11 AUX 11, 12

13 AUX 13, 14 (CMH, CML)

For example, if you specify the SMU input 1 for the couple port and set the couple mode ON, the input 1 will be connected to an odd number of the output port and the input 2 will connected to the next number of the output port. And if you specify to connect the input 1 to the output 10, the input 1 will be connected to the output 9 and the input 2 will be connected to the output 10.

If the ground input port and a couple port have been assigned to the same input port, the ground mode and the couple mode cannot be used in parallel.

If a ground enabled input port and a couple port have been assigned to the same input port, the ground mode and the couple mode cannot be used in parallel.

Page 81 of the Agilent Technologies B2201A user manual — Couple Mode
Page 81 · click to enlarge
Page 81

Front Panel Operation Switch Control Functions

The Agilent B2200 provides the function to detect the input ports that connect the Kelvin triaxial cable and set the ports as the couple port automatically. See “SCAN” on page 3-32

Page 82 of the Agilent Technologies B2201A user manual — LED Matrix
Page 82 · click to enlarge
Page 82

Front Panel Operation Display Functions

Display Functions

The Agilent B2200 provides LED matrix, LCD, and 18 front panel keys for front panel operation. This section explains the display of the LED matrix and the LCD.

  • “LED Matrix”
  • “LCD”

LED Matrix

Agilent B2200 has four blocks of 14 × 12 LED matrix. They display the switch condition of the switch modules installed in the slot 1 to slot 4. See Table 3-3.

Also LEDs labeled Card 1 to Card 4 are located above the LED matrices. They indicate the status of the module installed in the slot 1 to 4, respectively. See Table 3-4.

Table 3-3 LED Matrix

Green The switch has been closed to connect the output channel to the input port.

Red The switch has been closed to connect the output channel to the input bias port or the input ground port.

Orange The LED also blinks. The switch condition can be changed by the

Open/Close key.

- The LED lights out. The switch has been opened.

Table 3-4 Card 1/2/3/4 LED

Red The switch module has failed the selftest or the diagnostics.

Green The switch module has operated normally.

- The LED lights out. No switch module has been installed in this slot.

LED color Explanation · LED color Explanation

Page 83 of the Agilent Technologies B2201A user manual — A FREE BBM C C G - -
Page 83 · click to enlarge
Page 83

LCD

The LCD displays the setup information and the status information as shown in Table 3-5. Display example is shown in Figure 3-2.

Figure 3-2 LCD Display Example

01 02 03 04 05 06 07 08 09 10 11 12 13 14 Shift

A FREE BBM C C G - -

Sequence Rule Gnd Lock Error Gnd Bias

The first line shows the input port status, couple port, bias port, ground port, or ground enabled port.

The second line shows the instrument status and the mode status.

The example of Figure 3-2 notifies you the following information.

  • Input 5 and 6 are a couple port.
  • Input 12 is the input ground port.
  • Input 4 and 7 are the ground enabled input port.
  • Channel configuration mode is auto.
  • Couple mode is ON.
  • Ground mode is ON.
  • Connection rule is free.
  • Connection sequence is break before make.

Front Panel Operation · Display Functions

Page 84 of the Agilent Technologies B2201A user manual — LCD
Page 84 · click to enlarge
Page 84

Front Panel Operation Display Functions

Table 3-5 LCD Display Items

01 to 14 Port function assigned to the input ports 1 to 14. B (bias port), C (couple port), G (ground port), - (ground enabled port), or blank (no function).

Shift Shift key status indicator. The triangle mark appears when the sub key is active.

Remote Remote status indicator. The triangle mark appears when the Agilent B2200 is in the GPIB remote condition.

Lock Key status indicator. The triangle mark appears when the front panel keys are locked by an external computer.

Error Error status indicator. The triangle mark appears when any error has occurred.

Card# Slot number of the switch module to be monitored. Displays A when the channel configuration mode is Auto. In the auto mode, the modules installed in the mainframe from slot number 1 continuously will be controlled as one module. Or displays 1 to 4 when the channel configuration mode is Normal. The module can be selected by the Card Selection key.

Bias Bias mode status indicator. The triangle mark appears when the mode is ON.

Couple Couple mode status indicator. The triangle mark appears when the mode is ON.

Gnd Ground mode status indicator. The triangle mark appears when the mode is ON.

Rule Connection rule, FREE (plural route mode) or SINGLE (single route mode).

Sequence Connection sequence, BBM (Break Before Make), MBBR (Make Before Break), or NO_SEQ (No Sequence).

Label Description

Page 85 of the Agilent Technologies B2201A user manual — NO
Page 85 · click to enlarge
Page 85
Agilent Technologies B2201A — Front Panel Keys diagram (page 85)
Fig. 10 · Front Panel Keys

Front Panel Keys

The Agilent B2200 provides LED matrix, LCD, and 18 front panel keys for front panel operation. The front panel keys are used to change the instrument settings, the switch conditions, and so on.

Agilent Technologies B2201A — Front Panel Keys diagram (page 85)
Fig. 11 · Front Panel Keys

Local Sets the Agilent B2200 to the local condition.

Used to reset the Agilent B2200.

Displays the following message. Press the arrow key to set YES, and then press the

Enter key to reset the instrument.

NO

Menu Displays the setup menu. See “Setup Menu” on page 3-30.

Displays the selftest menu. See “Selftest Menu” on page 3-34.

Reset (Shift+Local) · Selftest (Shift+Menu)

Page 86 of the Agilent Technologies B2201A user manual — ENABLE DISABLE
Page 86 · click to enlarge
Page 86

Front Panel Operation Front Panel Keys

Function key group

The Function key group is used to change the input ports and modes.

Agilent Technologies B2201A — Function key group diagram (page 86)
Fig. 12 · Function key group

Port Function Enters into the port function selection mode. In this mode, the arrow keys, Fast key,

Exit key, and Enter key are available. Other keys will work to exit this mode.

In this mode, move cursor right or left to specify the input port, and press the up or down arrow key to select the port function, B (bias port), C (couple port), G (ground port), or - (ground enable port). The port function can be selected when the corresponding port operation mode (bias mode, couple mode, or ground mode) is set to ON. The bias mode and the ground mode cannot be set to ON simultaneously.

Bias Ch (Shift+ Port Function)

Displays the following menu.

SET VIEW

  • SET displays the following menu. The following functions are used to set the bias-enabled or bias-disabled channel (output port).

ENABLE DISABLE

The function displays the following message for example. And then press Enter to bias-enable or bias-disable the channel.

Channel01

If there is no enabled or disabled channel, the following message will appear.

ChannelNo Channel
  • VIEW displays the following message for example, and is used to see the bias status of the channel.
Channel 01ENABLED
Page 87 of the Agilent Technologies B2201A user manual — NO
Page 87 · click to enlarge
Page 87

Bias Mode Sets the bias mode ON or OFF.

When the bias mode is set to ON, the LCD shows which input port is the bias port, and the input bias port will be connected to the bias-enabled channels that are not connected to other input port. And LEDs will be turned red to show the bias port connections.

Ch Mode (Shift+ Bias Mode)

Sets the channel configuration mode AUTO or NORMAL.

Changing the mode will reset the Agilent B2200 settings.

Couple Mode Sets the couple mode ON or OFF.

When the couple mode is set to ON, the LCD shows which input ports are the couple port.

Gnd Mode (Shift+ Couple Mode)

Sets the ground mode ON or OFF.

When the ground mode is set to ON, the LCD shows which input port is the ground port, and the input ground port will be connected to the ground-enabled channels that are not connected to other input port. And LEDs will be turned red to show the ground port connections. The ground enabled input ports are also connected to the input ground port. Nothing must be connected to the ground enabled input ports.

Rule Displays the following message. Selects the connection rule FREE (plural route connections) or SINGLE (single route connection).

FREE

Displays the following message. Selects the connection sequence Break_Before_Make, Make_Before_Break, or No_Sequence.

Load Memory Displays the following message. Used to select the Agilent B2200 setup data stored in the internal memory, and loads the data.

Settings01 [ ]

Save Memory (Shift+ Load Memory)

Displays the following message. Used to select the internal memory to store the Agilent B2200 setup data, and stores the data. Up to eight setup data can be saved.

Settings01 [ ]

Open All Displays the following message. Press the arrow key to set YES, and then press the

Enter key to open all relays.

NO

Show 4 more labels / short text on this page
Front Panel Operation · Front Panel Keys · Sequence (Shift+Rule) · Break_Before_Make
Page 88 of the Agilent Technologies B2201A user manual — ENABLE DISABLE
Page 88 · click to enlarge
Page 88

Front Panel Operation Front Panel Keys

Gnd Ch (Shift+Open All)

Displays the following menu.

SET VIEW

  • SET displays the following menu. The following functions are used to set the ground-enabled or ground-disabled channel (output port).

ENABLE DISABLE

The function displays the following message for example. And then press Enter to ground-enable or ground-disable the channel.

Channel01

If there is no enabled or disabled channel, the following message will appear.

ChannelNo Channel
  • VIEW displays the following message for example, and is used to see the ground status of the channel.
Channel 01ENABLED
Page 89 of the Agilent Technologies B2201A user manual — Edit key group
Page 89 · click to enlarge
Page 89

Edit key group

The Edit key group is used to select the switch module to be monitored, move the cursor in the display, make the setup value effective, return to the previous menu or display, and so on.

Agilent Technologies B2201A — Edit key group diagram (page 89)
Fig. 13 · Edit key group

Card Selection Selects the switch module to be monitored on the LCD. The Card# status indicator will change the slot number (card number) when the Card Selection key is pressed.

This key is not effective when the Card# indicator shows A (auto configuration mode).

Shift Enables the sub key function. The sub key label has been printed on the front panel above the key. The label color is blue.

Exit Returns the display to the previous menu or display without making the setup effective.

Enter Makes the setup effective, and returns the display to the previous menu or display.

Fast Used with the arrow key. Pressing the Fast key and an arrow key speeds up changing the setting value.

arrow keys Moves the cursor on the display, displays another selection for a setup message, or changes the value of the digit specified by the cursor.

For the LED matrix, moves the blink-LED.

Open/Close When no blink-LED is on the LED matrix, enters into the LED matrix based switch control mode. The blink-LED will appear. In this mode, the arrow keys and Fast key are available. Other keys will work to exit this mode.

When the blink-LED is on the LED matrix, changes the condition, open or close, of the switch specified by the blink-LED.

Front Panel Operation · Front Panel Keys

Page 90 of the Agilent Technologies B2201A user manual — Setup Menu
Page 90 · click to enlarge
Page 90

Front Panel Operation Setup Menus

Setup Menus

This section explains the following setup menus displayed by pressing the Menu key or Selftest key.

  • “Setup Menu”
  • “Selftest Menu”

Setup Menu

Press the Menu key. The setup menu will be displayed on the LCD as shown in the following example. The setup menu is used to change the GPIB address, detect the couple port automatically, and so on.

[Hardware Configuration] CONFIG SCAN ERROR RMT_DSPL BEEP >

[Hardware Configuration]

Press the arrow key to move the cursor to the desired function name. Then, press the

Enter key. The setup message or the sub menu will be displayed. To return to the previous menu, press the Exit key.

Page 91 of the Agilent Technologies B2201A user manual — Setup Menus
Page 91 · click to enlarge
Page 91

Function Tree The setup menu provides the following functions.

  • CONFIG
  • ADDRESS Sets the GPIB address.
  • REVISION Displays the firmware revision.
  • UNIT Displays the module information of each slot.
  • SCAN Detects and sets the input couple ports automatically.
  • ERROR
  • DISPLAY Displays the error code and error message.
  • CLEAR Clears the error buffer.
  • RMT_DSPL
  • LCD Selects on or off for the LCD display in the remote mode.
  • LED Selects on or off for the LED matrix in the remote mode.
  • BEEP Selects on or off for the beeper.
  • PEN Enables or disables the light pen.

Front Panel Operation · Setup Menus

Page 92 of the Agilent Technologies B2201A user manual — NO
Page 92 · click to enlarge
Page 92

Front Panel Operation Setup Menus

CONFIG Displays the sub menu that provides the following functions.

  • ADDRESS

Displays the following message.

GPIB Address = Address

Press the arrow key to set the desired GPIB address. Then, press the Enter key to make the setup effective, or press the Exit key to cancel changing the setup.

  • REVISION

Displays the following message.

X.XX.XX

where X.XX.XX shows the firmware revision.

To return to the previous menu, press the Exit key.

  • UNIT

Displays the following message.

Slot#model , X

where # is 1, 2, 3, or 4, model is the model number of the module, X is the revision of the module.

To display another message, press the arrow key.

To return to the previous menu, press the Exit key.

SCAN Displays the following message.

NO

Press the arrow key to set YES. Then, press the Enter key to detect and set the input couple ports automatically, or press the Exit key to cancel this operation. The input couple ports will be the input ports the Kelvin cable is connected.

Page 93 of the Agilent Technologies B2201A user manual — NO
Page 93 · click to enlarge
Page 93

ERROR Displays the sub menu that provides the following functions.

  • DISPLAY

Displays the error code and error message, or No Error.

To return to the previous menu, press the Exit key.

  • CLEAR

Displays the following message. Press the arrow key to set YES. Then, press the

Enter key to clear the error buffer, or press the Exit key to cancel this operation.

NO

RMT_DSPL Displays the sub menu that provides the following functions.

  • LCD

Displays the following message. Press the arrow key to set ON (displays all status information in the GPIB remote mode) or OFF (displays Remote, Lock, and Error only). Then, press the Enter key to make the setup effective, or press the Exit key to cancel changing the setup.

  • LED

Displays the following message. Press the arrow key to set ON (enables LED matrix in the GPIB remote mode) or OFF (disables it). Then, press the Enter key to make the setup effective, or press the Exit key to cancel changing the setup.

BEEP Displays the following message. Press the arrow key to set the beeper ON or OFF. Then, press the Enter key to make the setup effective, or press the Exit key to cancel changing the setup.

PEN Displays the following message. Press the arrow key to enable or disable the light pen. Then, press the Enter key to make the setup effective, or press the Exit key to cancel changing the setup.

Show 6 more labels / short text on this page
Front Panel Operation · Setup Menus · OFF or ON · ON or OFF · ON or OFF · DISABLE or ENABLE
Page 94 of the Agilent Technologies B2201A user manual — Selftest Menu
Page 94 · click to enlarge
Page 94

Front Panel Operation Setup Menus

Selftest Menu

Press the Shift key and the Menu key. The selftest menu is displayed. The menu is used to execute the selftest and diagnostics.

[Execute Diagnostics] SELF_TEST RELAY_TEST KEY BEEPER >

[Execute Diagnostics]

Press the arrow key to move the cursor to the desired function name. Then, press the

Enter key. The setup message or the sub menu will be displayed. To return to the previous menu, press the Exit key.

Function Tree This menu provides the following functions.

  • SELFTEST Executes the controller test or displays test result.
  • RELAY_TEST Executes the relay test or displays test result.
  • KEY Executes the front panel interface test or displays test result.
  • BEEPER Executes the beeper test or displays test result.
  • LED Executes the LED matrix test or displays test result.
  • PEN Executes the light pen test or displays test result.
  • GPIB Executes the GPIB test or displays test result.
Page 95 of the Agilent Technologies B2201A user manual — Selftest Menu
Page 95 · click to enlarge
Page 95

SELF_TEST Displays the sub menu that provides the following functions.

  • EXECUTE

Starts the controller test. Wait until PASS or FAIL is displayed. Then press the

Exit key to return to the previous menu.

  • RESULT

Displays the test result. Press the Exit key to return to the previous menu.

RELAY_TEST Displays the sub menu that provides the following functions.

Before starting the relay test, disconnect cables from the input connectors. And open the outputs (end of cable is OK). For the Kelvin cable, disconnect it from the output connectors.

  • EXECUTE

Displays the following message.

Press the up or down arrow key to select the module to be tested. Then press the

Enter key to start the relay test, or press the Exit key to cancel this operation.

Number of modules is 2 or more, the relay test will be automatically completed. However, if the only one module is installed, the following message will appear.

Connect IV1 and 2. Press [Enter].

Then connect a cable between the input 1 and 2, and press Enter key.

After that, change the cable connection as shown in the displayed message and repeat this for the other input connections (input 3 and 4, 5 and 6, 7 and 8, 9 and 10, 11 and 12, 13 and 14). Do not forget to open the connectors other than the specified pair.

  • RESULT

Displays the following message (example). Press the up or down arrow key to display the test result for each module. Press the Exit key to return to the previous menu.

Slot1Not performed yet

Front Panel Operation · Setup Menus · Slot1

Page 96 of the Agilent Technologies B2201A user manual — Selftest Menu
Page 96 · click to enlarge
Page 96

Front Panel Operation Setup Menus

KEY Displays the sub menu that provides the following functions.

  • EXECUTE

Displays the following message.

DIAG:KEY Press any key.

Press any front panel key and confirm the LCD display. For example, if you press the Port Function key, the following message must be displayed.

DIAG:KEY [Port Function]

Continue this for all keys. After that, press the Enter key twice for the normal operation. The test result will be PASS.

If you find any wrong response, press the Exit key twice. The test result will be FAIL.

Finally, press the Exit key to return to the previous menu.

  • RESULT

Displays the test result. Press the Exit key to return to the previous menu.

BEEPER Displays the sub menu that provides the following functions.

  • EXECUTE

Starts the beeper test, and displays the following message.

DIAG:BEEP Is beeper making 2 sounds?

Press the Enter key if the beeper works correctly. The test result will be recorded as PASS.

If you find any problem, press the Exit key. The test result will be recorded as FAIL.

Finally, press the Exit key to return to the previous menu.

  • RESULT

Displays the test result. Press the Exit key to return to the previous menu.

Page 97 of the Agilent Technologies B2201A user manual — SLOT: 0 INPUT: 5 OUTPUT:10
Page 97 · click to enlarge
Page 97

LED Displays the sub menu that provides the following functions.

  • EXECUTE

Starts the LED matrix test, and displays the following message.

DIAG:LED All LED in Orange?

Press the Enter key if the all LEDs are lighted in orange. The test result will be recorded as PASS.

If you find any problem, press the Exit key. The test result will be recorded as FAIL.

Finally, press the Exit key to return to the previous menu.

  • RESULT

Displays the test result. Press the Exit key to return to the previous menu.

PEN Displays the sub menu that provides the following functions.

Before starting the light pen test, connect the light pen to the Agilent B2200.

  • EXECUTE

Starts the light pen test, and displays the following message.

DIAG:PEN Point any cross point

Point the light pen to the point you desire and confirm the LCD display. For example, if you point the input 5-to-output 10 cross point of the Agilent B2200 in the auto configuration mode, the following message must be displayed.

SLOT: 0 INPUT: 5 OUTPUT:10

Continue this for all cross points. After that, press the Enter key for the normal operation. The test result will be recorded as PASS.

If you find any wrong response, press the Exit key. The test result will be recorded as FAIL.

Finally, press the Exit key to return to the previous menu.

  • RESULT

Displays the test result. Press the Exit key to return to the previous menu.

Front Panel Operation · Setup Menus

Page 98 of the Agilent Technologies B2201A user manual — Selftest Menu
Page 98 · click to enlarge
Page 98

Front Panel Operation Setup Menus

GPIB Displays the sub menu that provides the following functions.

  • EXECUTE

Displays the following message.

Open GPIB, then press [Enter]

Disconnect the cable from the GPIB connector on the rear panel, and open it.

Press the Enter key to start the GPIB test, and wait until PASS or FAIL is displayed. Then press the Exit key to return to the previous menu.

  • RESULT

Displays the test result. Press the Exit key to return to the previous menu.

Page 99 of the Agilent Technologies B2201A user manual — 4 Programming
Page 99 · click to enlarge
Page 99

4 Programming

Page 100 of the Agilent Technologies B2201A user manual — 4 Programming
Page 100 · click to enlarge
Page 100

This chapter describes the automatic control programming of the Agilent B2200.

  • “Programming Basics”

Explains the commands for the fundamental switch control of the Agilent B2200.

  • “Programming Examples”

Provides some examples of the programming.

  • “Capacitance Compensation”

Explains how to use the capacitance compensation routine. The routine is a function of the Agilent B2200 VXIplug&play driver.

For the details about SCPI commands, see Chapter 5.

For the details about VXIplug&play driver, see Chapter 6.

Programming

Page 101 of the Agilent Technologies B2201A user manual — :ROUT:CONN:RULE ALL,FREE
Page 101 · click to enlarge
Page 101

Programming Basics

This section explains the commands used for the fundamental switch control of the Agilent B2200.

  • “SCPI Command Hierarchy”
  • “Fundamental Commands”
  • “Switch Control”

SCPI Command Hierarchy

The SCPI commands use a hierarchical structure for subsystem commands similar to a file system. For example, in :ROUT:CONN:RULE command, the hierarchy is as follows:

The colon at the beginning of the command means root. The colons between two command keywords means moving down to a lower level.

A semicolon enables two commands to be sent on the same line. For example, :ROUT:CONN:RULE ALL,FREE;SEQ ALL,BBM is the same as the following two commands:

:ROUT:CONN:RULE ALL,FREE

:ROUT:CONN:SEQ ALL,BBM

So, using a semicolon reduces typing and simplifies the program. A command terminator (such as carriage return) resets the path to root.

Show 4 more labels / short text on this page
Programming Programming Basics · ROUT root · CONN sub-level 1 · RULE sub-level 2
Page 102 of the Agilent Technologies B2201A user manual — :ROUT:FUNC NCON
Page 102 · click to enlarge
Page 102

Fundamental Commands

The following commands are used to set the fundamental switch control functions of the Agilent B2200. The commands should be entered before performing the open/close operation. For the functions, see “Switch Control Functions” on page 3-12.

Table 4-1 Fundamental Commands

Sets the channel configuration mode.

:ROUT:FUNC NCON

:ROUT:FUNC ACON

Sets the connection rule. :ROUT:CONN:RULE card_no,FREE

:ROUT:CONN:RULE card_no,SROU

Sets the connection sequence. :ROUT:CONN:SEQ card_no,NSEQ

:ROUT:CONN:SEQ card_no,BBM

:ROUT:CONN:SEQ card_no,MBBR

Programming Programming Basics · Functions Commands

Page 103 of the Agilent Technologies B2201A user manual — ,
Page 103 · click to enlarge
Page 103

Switch Control

The following commands are used to control open/close of the specified switch.

Table 4-2 Switch Control Commands

Closes the relays specified by channel_list. :ROUT:CLOS channel_list

Opens the relays specified by channel_list. :ROUT:OPEN channel_list

Opens the all relays on the card specified by card_no.

:ROUT:OPEN:CARD card_no

The channel_list is the parameter which determines the input-output cross points to open/close. Figure 4-1 shows the basic syntax of the channel_list parameter.

In the channel_list, you can specify one or more channels. Each specified channel consists of five digits.

Card No.: One digit. 0 when the channel configuration mode is AUTO. Or the slot number (1 to 4) that installs the switch module when the mode is NORMAL.

Channel No.: First two digits: specifies input port. 01 to 14.

Last two digits: specifies output port. 01 to 12, 01 to 24, 01 to 36, or 01 to 48 for the AUTO mode. It depends on the number of modules installed in the mainframe from slot number 1 continuously. Or 01 to 12 for the NORMAL mode

See Figure 1-4.

Figure 4-1 Syntax of channel_list Parameter

,

,

Channel No. (4 digits) Card No. (1digit) @ Channel No. (4 digits) Card No. (1digit) ( :

) Channel No. (4 digits) Card No. (1digit) @ Channel No. (4 digits) Card No. (1digit) ( :

Programming Programming Basics · Functions Commands · )

Page 104 of the Agilent Technologies B2201A user manual — Switch Control
Page 104 · click to enlarge
Page 104

In the channel_list, you can specify multiple channels by using comma (,) or colon (:) as follows:

Commause between each specified channel as in following examples:
  • (@10101,10102,10103) means 10101, 10102 and 10103.
  • (@10112,10202) means 10112 and 10202.
  • (@11412,20102) means 11412 and 20102.
Colonuse to specify a range of channels as in following examples:
  • (@10101:10103) means 10101, 10102, and 10103.
  • (@10112:10202) means 10112, 10201, and 10202.
  • (@11412:20102) means 11412, 20101, and 20102.

As shown in last two examples, you can specify a range across input ports and card slots.

In the Auto Configuration mode, you can omit any zeros at beginning of channel. For example, channel 00101 can be expressed by 101.

For the Kelvin outputs, the input-output paths must be controlled as shown below.

  • For the input couple port:

Control the switches to connect the input couple port to the Kelvin output.

  • For the single input port:

Control the switches to connect the input port to two ports of the Kelvin output.

If you do not comply with this note, leakage current between the Kelvin paths will cause measurement error.

Programming Programming Basics

Page 105 of the Agilent Technologies B2201A user manual — Programming Examples
Page 105 · click to enlarge
Page 105

Programming Examples

This section provides examples of control programs for the Agilent B2200.

  • “Connecting Input-Output Paths”
  • “Using Bias Mode”
  • “Using Ground Mode”
  • “Using Couple Mode”
  • “Saving Input/Output Labels”
  • “Defining Comment for Internal Memory”

Example programs use the Microsoft Visual Basic .NET and the Agilent T&M Programmers Toolkit.

The program examples do not include instrument control routines, so to make a measurement, you need to add the desired routines.

If you assign the input port 12 to the ground port, the input port 12 will be connected to the ground internally when the ground mode is ON. In this case, leave open the input port 12.

Programming Programming Examples

Page 106 of the Agilent Technologies B2201A user manual — Connecting Input-Output Paths
Page 106 · click to enlarge
Page 106

Connecting Input-Output Paths

The following example connects instrument output to DUT as shown in Figure 4-2.

Figure 4-2 Input-Output Connection Example

Agilent Technologies B2201A — Connecting Input-Output Paths diagram (page 106)
Fig. 14 · Connecting Input-Output Paths
Setup• Channel configuration mode: Normal
  • Connection rule: Single
  • Connection sequence: Break_Before_Make
  • Display strings: “Connecting MOSFET AG002201”
  • Used module: Switch module installed in the slot 1.
  • Connection paths:

from SMU1 to Output 1 (channel list 10101)

from SMU2 to Output 2 (channel list 10202)

from SMU3 to Output 3 (channel list 10303)

from SMU4 to Output 4 (channel list 10404)

Programming Programming Examples

Page 107 of the Agilent Technologies B2201A user manual — Connecting Input-Output Paths
Page 107 · click to enlarge
Page 107

Table 4-3 Input-Output Connection Example

Imports Agilent.TMFramework '1 Imports Agilent.TMFramework.InstrumentIO

Dim B2200 As New DirectIO("GPIB0::22::INSTR") '6 Dim channels As String = "(@10101,10202,10303,10404)" B2200.WriteLine("*RST") '8 B2200.WriteLine(":ROUT:FUNC NCON") B2200.WriteLine(":ROUT:CONN:RULE ALL,SROU") B2200.WriteLine(":ROUT:CONN:SEQ ALL,BBM") B2200.WriteLine(":SYST:DISP:STR 'Connecting MOSFET AG002201'") B2200.WriteLine(":ROUT:CLOS " & channels) '13 MsgBox("Click OK to start measurement.", vbOKOnly, "") Console.WriteLine("Measurement in progress. . ." & Chr(10)) 'insert the code for measurement

B2200.WriteLine(":ROUT:OPEN:CARD ALL") '17 B2200.Close() MsgBox("Click OK to stop the program.", vbOKOnly, "") Console.WriteLine("Measurement completed." & Chr(10)) End Sub '21 End Module

1 to 6 The above example is for the B2200 of the GPIB address 22 on the interface GPIB0. “GPIB0” is the VISA name. Confirm your GPIB settings, and set them properly.

7 Defines the channel list in the channels variable (string).

8 to 12 Resets the Agilent B2200, sets the Normal configuration mode, the Single connection rule, and the Break_Before_Make connection sequence. And displays the string on the LCD.

13 Closes switches to connect the input-output paths specified by channels.

14 to 16 Opens a message box to confirm the start of measurement. And clicking OK on the message box displays a message on the console window.

Insert measurement control program code at the line 16.

17 Opens all relays on the all modules installed in the Agilent B2200.

18 to 22 Disables the software connection with the Agilent B2200, and opens a message box to confirm the end of the program. Finally, clicking OK on the message box displays a message on the console window.

Show 4 more labels / short text on this page
Programming Programming Examples · Module Module1 · Sub Main() · Line Description
Page 108 of the Agilent Technologies B2201A user manual — Using Bias Mode
Page 108 · click to enlarge
Page 108

Using Bias Mode

The following example uses the bias mode.

Figure 4-3 Bias Mode Example

Setup• Channel configuration mode: Auto
  • Connection rule: Single. Multiple connection is available for the bias port.
  • Connection sequence: Break_Before_Make
  • Used module: All switch modules installed in the mainframe.
  • Bias port: AUX Input 10
  • Bias enabled output port: All output ports
  • Connection paths (connected when the bias mode is ON):

from bias port to all bias enabled output ports

  • Connection paths (connected by the :ROUT:CLOS command):

The following output ports will be disconnected from the bias port, and the following input-output connections will be made.

from SMU1 to Output 1 (channel list 00101)

from SMU2 to Output 2 (channel list 00202)

from SMU3 to Output 3 (channel list 00303)

from SMU4 to Output 4 (channel list 00404)

Programming Programming Examples

Page 109 of the Agilent Technologies B2201A user manual — Using Bias Mode
Page 109 · click to enlarge
Page 109

Table 4-4 Bias Mode Example

Imports Agilent.TMFramework Imports Agilent.TMFramework.DataAnalysis Imports Agilent.TMFramework.DataVisualization Imports Agilent.TMFramework.InstrumentIO Module Module1

Dim B2200 As New DirectIO("GPIB0::22::INSTR") Dim channels As String = "(@101,202,303,404)" '8 B2200.WriteLine("*RST") B2200.WriteLine(":ROUT:FUNC ACON") '10 B2200.WriteLine(":ROUT:CONN:RULE ALL,SROU") B2200.WriteLine(":ROUT:CONN:SEQ ALL,BBM") B2200.WriteLine(":ROUT:BIAS:PORT ALL,10") B2200.WriteLine(":ROUT:BIAS:CHAN:ENAB:CARD ALL") B2200.WriteLine(":ROUT:BIAS:STAT ALL,ON") MsgBox("Click OK to start stress output.", vbOKOnly, "") '16 'insert the code for stress output

B2200.WriteLine(":ROUT:BIAS:STAT ALL,OFF") B2200.WriteLine(":ROUT:CLOS " & channels) '19 MsgBox("Click OK to start measurement.", vbOKOnly, "") Console.WriteLine("Measurement in progress. . ." & Chr(10)) 'insert the code for measurement

B2200.WriteLine(":ROUT:OPEN:CARD ALL") B2200.Close() MsgBox("Click OK to stop the program.", vbOKOnly, "") Console.WriteLine("Measurement completed." & Chr(10)) End Sub '27 End Module

8 Defines the channel list in the channels variable (string).

9 to 12 Resets the Agilent B2200. And sets the Auto configuration mode, the Single connection rule, and the Break_Before_Make connection sequence.

13 to 15 Sets the bias port and the bias enabled output ports. And turns the bias mode ON.

16 to 17 Opens a message box to confirm the start of measurement. And clicking OK on the message box displays a message on the console window.

Insert bias source control program code at the line 17.

18 Turns the bias mode OFF.

19 to 28 Same as the code shown in the lines 13 to 22 of Table 4-3.

Programming Programming Examples · Sub Main() · Line Description

Page 110 of the Agilent Technologies B2201A user manual — Using Ground Mode
Page 110 · click to enlarge
Page 110

Using Ground Mode

The following example uses the ground mode.

Figure 4-4 Ground Mode Example

Agilent Technologies B2201A — Using Ground Mode diagram (page 110)
Fig. 15 · Using Ground Mode
Setup• Channel configuration mode: Auto
  • Connection rule: Single. Multiple connection is available for the ground port.
  • Connection sequence: Break_Before_Make
  • Used module: All switch modules installed in the mainframe.
  • Ground port: AUX Input 12
  • Ground enabled input port: SMU Input 5 to 8
  • Ground enabled output port: All output ports
  • Connection paths (connected when the ground mode is ON):

from ground port to all ground enabled input/output ports

  • Connection paths (connected by the :ROUT:CLOS command):

The following output ports will be disconnected from the ground port, and the following input-output connections will be made.

from SMU1 to Output 1 (channel list 00101)

from SMU2 to Output 2 (channel list 00202)

from SMU3 to Output 3 (channel list 00303)

from SMU4 to Output 4 (channel list 00404)

Programming Programming Examples

Page 111 of the Agilent Technologies B2201A user manual — Using Ground Mode
Page 111 · click to enlarge
Page 111

Table 4-5 Ground Mode Example

Imports Agilent.TMFramework Imports Agilent.TMFramework.DataAnalysis Imports Agilent.TMFramework.DataVisualization Imports Agilent.TMFramework.InstrumentIO Module Module1

Dim B2200 As New DirectIO("GPIB0::22::INSTR") Dim channels As String = "(@101,202,303,404)" '8 B2200.WriteLine("*RST") B2200.WriteLine(":ROUT:FUNC ACON") '10 B2200.WriteLine(":ROUT:CONN:RULE ALL,SROU") B2200.WriteLine(":ROUT:CONN:SEQ ALL,BBM") B2200.WriteLine(":ROUT:AGND:PORT ALL,12") '13 B2200.WriteLine(":ROUT:AGND:UNUSED ALL,'5,6,7,8'") B2200.WriteLine(":ROUT:AGND:CHAN:ENAB:CARD ALL") B2200.WriteLine(":ROUT:AGND:STAT ALL,ON") B2200.WriteLine(":ROUT:CLOS " & channels) '17 MsgBox("Click OK to start measurement.", vbOKOnly, "") Console.WriteLine("Measurement in progress. . ." & Chr(10)) 'insert the code for measurement

B2200.WriteLine(":ROUT:AGND:STAT ALL,OFF") '21 B2200.WriteLine(":ROUT:OPEN:CARD ALL") B2200.Close() MsgBox("Click OK to stop the program.", vbOKOnly, "") Console.WriteLine("Measurement completed." & Chr(10)) End Sub '26 End Module

8 Defines the channel list in the channels variable (string).

9 to 12 Resets the Agilent B2200. And sets the Auto configuration mode, the Single connection rule, and the Break_Before_Make connection sequence.

13 to 16 Sets the ground port and the ground enabled input/output ports. And turns the ground mode ON. Open the ground enabled input ports to prevent the instruments from damage.

17 Closes switches to connect the input-output paths specified by channels.

17 to 27 Almost same as the code shown in the lines 13 to 22 of Table 4-3. The command used to turn the ground mode OFF has been inserted to the line 21.

Programming Programming Examples · Sub Main() · Line Description

Page 112 of the Agilent Technologies B2201A user manual — Using Couple Mode
Page 112 · click to enlarge
Page 112

Using Couple Mode

The following example uses the couple mode.

Figure 4-5 Couple Mode Example

Agilent Technologies B2201A — Using Couple Mode diagram (page 112)
Fig. 16 · Using Couple Mode
Setup• Channel configuration mode: Auto
  • Connection rule: Single
  • Connection sequence: Break_Before_Make
  • Used module: All switch modules installed in the mainframe.
  • Couple port: SMU Input 3-4 and 5-6
  • Connection paths:

from SMU1 to Output 1 (channel list 00101)

from SMU2 to Output 2 (channel list 00202)

from SMU3 to Output 3 (channel list 00303)

from SMU4 to Output 4 (coupled with the channel list 00303)

from SMU5 to Output 5 (channel list 00505)

from SMU6 to Output 6 (coupled with the channel list 00505)

Programming Programming Examples

Page 113 of the Agilent Technologies B2201A user manual — Using Couple Mode
Page 113 · click to enlarge
Page 113

Table 4-6 Couple Mode Example

Imports Agilent.TMFramework Imports Agilent.TMFramework.DataAnalysis Imports Agilent.TMFramework.DataVisualization Imports Agilent.TMFramework.InstrumentIO Module Module1

Dim B2200 As New DirectIO("GPIB0::22::INSTR") Dim channels As String = "(@101,202,303,505)" '8 B2200.WriteLine("*RST") B2200.WriteLine(":ROUT:FUNC ACON") '10 B2200.WriteLine(":ROUT:CONN:RULE ALL,SROU") B2200.WriteLine(":ROUT:CONN:SEQ ALL,BBM") B2200.WriteLine(":ROUT:COUP:PORT ALL,'3,5'") '13 B2200.WriteLine(":ROUT:COUP:STAT ALL,ON") B2200.WriteLine(":ROUT:CLOS " & channels) '15 MsgBox("Click OK to start measurement.", vbOKOnly, "") Console.WriteLine("Measurement in progress. . ." & Chr(10)) 'insert the code for measurement

B2200.WriteLine(":ROUT:COUP:STAT ALL,OFF") '19 B2200.WriteLine(":ROUT:OPEN:CARD ALL") B2200.Close() MsgBox("Click OK to stop the program.", vbOKOnly, "") Console.WriteLine("Measurement completed." & Chr(10)) End Sub '24 End Module

8 Defines the channel list in the channels variable (string).

9 to 12 Resets the Agilent B2200. And sets the Auto configuration mode, the Single connection rule, and the Break_Before_Make connection sequence.

13 to 14 Sets the couple ports, and turns the couple mode ON.

15 Closes switches to connect the input-output paths specified by channels.

15 to 25 Almost same as the code shown in the lines 13 to 22 of Table 4-3. The command used to turn the couple mode OFF has been inserted to the line 19.

Programming Programming Examples · Sub Main() · Line Description

Page 114 of the Agilent Technologies B2201A user manual — Saving Input/Output Labels
Page 114 · click to enlarge
Page 114

Saving Input/Output Labels

You can define labels to use for the input/output ports when you control the switch in the GPIB local mode. The labels are cleared by *RST. So it is recommended to save the labels into the internal memory and define a comment for the memory data. You can see the comment when you load/save the memory data in the GPIB local mode.

Setup• Channel configuration mode: Auto
  • Label for input 1: SMU1
  • Label for input 2: SMU2
  • Label for input 3: SMU3
  • Label for input 4: UNUSED
  • Label for input 5: SMU4-F
  • Label for input 6: SMU4-S
  • Label for input 7: SMU5-F
  • Label for input 8: SMU5-S
  • Label for input 9: OPEN
  • Label for input 10: BIAS
  • Label for input 11: OPEN
  • Label for input 12: GROUND
  • Label for input 13: CMU-H
  • Label for input 14: CMU-L
  • Label for output 1: BULK
  • Label for output 2: SOURCE
  • Label for output 3: GATE
  • Label for output 4: DRAIN
  • Label for output 5: GROUND
  • Comment for memory 1: Port label info

Programming Programming Examples

Page 115 of the Agilent Technologies B2201A user manual — Saving Input/Output Labels
Page 115 · click to enlarge
Page 115

Table 4-7 Label Definition and Data Save Example

Imports Agilent.TMFramework Imports Agilent.TMFramework.InstrumentIO Module Module1

Dim B2200 As New DirectIO("GPIB0::22::INSTR") B2200.WriteLine("*RST") B2200.WriteLine(":ROUT:FUNC ACON") Console.WriteLine("Starts labeling." & Chr(10)) B2200.WriteLine(":SYST:DISP:STR 'Updating memory 1 data.'") B2200.WriteLine(":ROUT:SYMB:PORT 1,'SMU1 '") '10 B2200.WriteLine(":ROUT:SYMB:PORT 2,'SMU2 '") B2200.WriteLine(":ROUT:SYMB:PORT 3,'SMU3 '") B2200.WriteLine(":ROUT:SYMB:PORT 4,'UNUSED'") B2200.WriteLine(":ROUT:SYMB:PORT 5,'SMU4-F'") B2200.WriteLine(":ROUT:SYMB:PORT 6,'SMU4-S'") B2200.WriteLine(":ROUT:SYMB:PORT 7,'SMU5-F'") B2200.WriteLine(":ROUT:SYMB:PORT 8,'SMU5-S'") B2200.WriteLine(":ROUT:SYMB:PORT 9,'OPEN '") B2200.WriteLine(":ROUT:SYMB:PORT 10,'BIAS '") B2200.WriteLine(":ROUT:SYMB:PORT 11,'OPEN '") B2200.WriteLine(":ROUT:SYMB:PORT 12,'GROUND'") B2200.WriteLine(":ROUT:SYMB:PORT 13,'CMU-H '") B2200.WriteLine(":ROUT:SYMB:PORT 14,'CMU-L '") B2200.WriteLine(":ROUT:SYMB:CHAN ALL,1,'BULK '") B2200.WriteLine(":ROUT:SYMB:CHAN ALL,2,'SOURCE'") B2200.WriteLine(":ROUT:SYMB:CHAN ALL,3,'GATE '") B2200.WriteLine(":ROUT:SYMB:CHAN ALL,4,'DRAIN '") B2200.WriteLine(":ROUT:SYMB:CHAN ALL,5,'GROUND'") '28 B2200.WriteLine(":SYST:MEMO:SAVE 1") B2200.WriteLine(":SYST:MEMO:COMM 1,'Port label info '") B2200.WriteLine(":SYST:DISP:STR 'Memory 1 data was updated.'") B2200.Close() Console.WriteLine("Labeling completed." & Chr(10)) End Sub '34 End Module

9, 31 Displays message on the LCD.

10 to 28 Defines labels for the inputs 1 to 14 and the outputs 1 to 5.

29 Saves the switch module setup including the labels into the internal memory 1.

30 Defines the comment “Port label info” for the memory 1.

Programming Programming Examples · Sub Main() · Line Description

Page 116 of the Agilent Technologies B2201A user manual — Defining Comment for Internal Memory
Page 116 · click to enlarge
Page 116

Defining Comment for Internal Memory

You can define comments for the internal memories. You can see the comment when you load/save the memory data in the GPIB local mode.

Setup• Channel configuration mode: Auto
  • Comment for memory 1: 1-1,2-2,3-3,5-15
  • Comment for memory 2: 1-1,2-2,3-4,5-17
  • Comment for memory 3: 1-1,2-2,3-5,5-19
  • Comment for memory 4: 1-1,2-2,3-7,5-21
  • Comment for memory 5: 1-1,2-2,3-9,5-23
  • Comment for memory 6: 10-BIAS
  • Comment for memory 7: 12-GROUND
  • Comment for memory 8: CMH-10,CML-11

Programming Programming Examples

Page 117 of the Agilent Technologies B2201A user manual — Defining Comment for Internal Memory
Page 117 · click to enlarge
Page 117

Table 4-8 Memory Comment Definition Example

Imports Agilent.TMFramework Imports Agilent.TMFramework.DataAnalysis Imports Agilent.TMFramework.DataVisualization Imports Agilent.TMFramework.InstrumentIO Module Module1

Dim B2200 As New DirectIO("GPIB0::22::INSTR") B2200.WriteLine("*RST") '8 B2200.WriteLine(":ROUT:FUNC ACON") Console.WriteLine("Starts labeling." & Chr(10)) B2200.WriteLine(":SYST:DISP:STR 'Updating memory comment.'") B2200.WriteLine(":SYST:MEMO:COMM 1,'1-1,2-2,3-3,5-15'") '12 B2200.WriteLine(":SYST:MEMO:COMM 2,'1-1,2-2,3-4,5-17'") B2200.WriteLine(":SYST:MEMO:COMM 3,'1-1,2-2,3-5,5-19'") B2200.WriteLine(":SYST:MEMO:COMM 4,'1-1,2-2,5-7,7-21'") B2200.WriteLine(":SYST:MEMO:COMM 5,'1-1,2-2,5-9,7-23'") B2200.WriteLine(":SYST:MEMO:COMM 6,'10-BIAS '") B2200.WriteLine(":SYST:MEMO:COMM 7,'12-GROUND '") B2200.WriteLine(":SYST:MEMO:COMM 8,'CMH-10,CML-11 '") '19 B2200.WriteLine(":SYST:DISP:STR 'Memory comment was updated.'") B2200.Close() Console.WriteLine("Labeling completed." & Chr(10)) End Sub '23 End Module

8 Resets the Agilent B2200.

9 Sets the channel configuration mode to Auto.

12 to 19 Defines the comments for the internal memory 1 to 8.

Programming Programming Examples · Sub Main() · Line Description

Page 118 of the Agilent Technologies B2201A user manual — Capacitance Compensation Function
Page 118 · click to enlarge
Page 118

Capacitance Compensation

When the capacitance/conductance measurement is performed through the Agilent B2200, LCR meter measures the capacitance/conductance of the path including a device under test (DUT), matrix switches, extension cables and so on. So, the data measured by the LCR meter is far from the DUT’s capacitance/conductance.

The Agilent B2200 VXIplug&play driver provides the functions used to compensate the capacitance/conductance measured by the Agilent 4284A LCR meter in the measurement environments described in “Required Conditions” on page 4-21.

This section explains how to use the capacitance compensation function.

  • “Capacitance Compensation Function”
  • “Required Conditions”
  • “To Create Compensation Data File”
  • “To Perform Measurement and Compensation”

Capacitance Compensation Function

Driver functions used for the capacitance compensation are listed below.

  • agb220xa_selectCompenFile function
  • agb220xa_compenC function

Capacitance measurement accuracy (typical): ±1 % ±0.5 pF

This typical data is for the following measurement conditions:

Measurement frequency1 kHz to 1 MHz
Measurement rangeMaximum 1000 pF

Measurement terminal: At the end of the Agilent 16494A/B/C cable connected to the switch module output terminals.

The typical data does not apply to anything extended from the 16494A/B/C cable. The conditions described in “Required Conditions” on page 4-21 must be satisfied.

Programming Capacitance Compensation

Page 119 of the Agilent Technologies B2201A user manual — Required Conditions
Page 119 · click to enlarge
Page 119

Required Conditions

The following conditions must be satisfied to use the capacitance compensation function. For the instrument connections, see Figure 4-6.

  • Setting of the 4284A
  • Option required: 4284A-006
  • Range of the measurement frequency: 1 kHz to 1 MHz
  • Measurement function: Cp-G
  • Connection to Agilent B2200

Use the Agilent 16494F CMU cable or the Agilent 16048D/E test leads to connect between the Agilent 4284A and the Agilent B2200 inputs.

If the 16048D/E is used, the BNC-T adapters (2 ea., Agilent part number 1250-2405 for each) are required to connect between the Hc and Hp terminals and between the Lc and Lp terminals.

  • Calibration

Perform the 4284A open calibration at the end of the measurement paths in front of the B2200 inputs. If you also perform the short calibration (optional), prepare the BNC thru adapter (Agilent part number 1250-0080, 1 ea.).

  • Total cable length of both Hc-Hp side and Lc-Lp side must be the same.
  • Agilent B2200 input ports

AUX Input 13 (CMH, for 4284A Hc-Hp) and 14 (CML, for 4284A Lc-Lp)

  • Connection from the Agilent B2200 outputs to the connector plate or the Agilent B2220A probe card interface

Use the Agilent 16494A triaxial cable or Agilent 16494B/C Kelvin triaxial cable.

  • Ahead of the connector plate
Recommended cableAgilent part number 8121-1191 Triaxial cable

You can also use another type of triaxial cable, coaxial cable, or combination of these.

To approximate the capacitance/conductance of the DUT, you need to obtain the appropriate compensation coefficients for your measurement environment, and create your compensation data file. See Figure 4-6.

Programming Capacitance Compensation

Page 120 of the Agilent Technologies B2201A user manual — Required Conditions
Page 120 · click to enlarge
Page 120

In Figure 4-6, C2H, C2L, C3H, C3L are the compensation coefficients defined in the compensation data file. where, CxH is for the path connected to the Agilent 4284A Hc-Hp terminal, and CxL is for the path connected to the Agilent 4284A Lc-Lp terminal.

When the Agilent B2220A probe card interface is used, obtain the coefficients for C3x, and create your compensation data file. In this case, probe card will be used for the C3x path.

When the connector plate is used, obtain the coefficients for C2x and C3x, and create your compensation data file. In this case, triaxial cable with connector plate will be used for the C2x path. And coaxial cable with positioner will be used for the C3x path.

For obtaining the compensation coefficients and creating the compensation data file, see “To Create Compensation Data File” on page 4-23.

Figure 4-6 Extension Cables and Compensation Coefficients

Agilent Technologies B2201A — Required Conditions diagram (page 120)
Fig. 17 · Required Conditions

Programming Capacitance Compensation

Page 121 of the Agilent Technologies B2201A user manual — To Create Compensation Data File
Page 121 · click to enlarge
Page 121

To Create Compensation Data File

This section explains how to create the compensation data file.

1. Select one of the compensation data files (template, 20 files) installed when the Agilent B2200 VXI plug&play driver is installed.

To select the most appropriate template for your measurement environment, see Table 4-9 that lists the file name and the measurement environment where the template targets. Each template is a text file that contains the information as shown below.

# # Compensation data file for B2210A/Probecard I/F/3m triax cable # # # MB Mother Board # MH Matrix Path High # ML Matrix Path Low # C1 Agilent Triax Cable # C2H Probe card I/F or User Triax cable High # C2L Probe card I/F or User Triax cable Low # C3H Probe card or User Coax cable High # C3L Probe card or User Coax cable Low # format version 1.0 B2210A PCIF # # R[Ohm] L[H] C[F] # MB 0.000000e+00 5.250000e-08 2.940000e-11 MH 2.430000e+00 6.310000e-07 1.930000e-10 ML 2.490000e+00 5.970000e-07 1.920000e-10 C1 6.300000e-01 1.250000e-06 1.600000e-10 C2H 2.988000e-01 5.090000e-07 7.000000e-11 C2L 2.988000e-01 5.090000e-07 7.000000e-11 C3H 0.000000e+00 8.000000e-08 1.500000e-13 C3L 0.000000e+00 8.000000e-08 1.500000e-13

The 15th line specifies the used switch module, B2210A or B2211A.

The 16th line specifies the DUT interface, PCIF or CABLE. PCIF indicates that the Agilent B2220A probe card interface is used. CABLE indicates that the connector plate is used.

The lines C2H to C3L should be modified for each measurement environment. See Table 4-9 and Table 4-10. Do not modify the other lines.

Programming Capacitance Compensation

Page 122 of the Agilent Technologies B2201A user manual — To Create Compensation Data File
Page 122 · click to enlarge
Page 122

Table 4-9 Template Compensation Data Files

Measurement environment that template targets

Switch module Cableb

\B2210A\pcif\triax\3m.data B2210A 16494A-002 B2220A C3H and C3L

\B2210A\pcif\triax\4m.data 16494A-005

\B2210A\pcif\kelvin\3m.data 16494C-002

\B2210A\pcif\kelvin\4m.data 16494C-005

\B2210A\cable\triax\1_5m.data 16494A-001 16495F/G C2H, C2L, C3H, and C3L \B2210A\cable\triax\3m.data 16494A-002

\B2210A\cable\triax\4m.data 16494A-005

\B2210A\cable\kelvin\1_5m.data 16494B-001

\B2210A\cable\kelvin\3m.data 16494B-002

\B2210A\cable\kelvin\4m.data 16494C-005

\B2211A\pcif\triax\3m.data B2211A 16494A-002 B2220A C3H and C3L

\B2211A\pcif\triax\4m.data 16494A-005

\B2211A\pcif\kelvin\3m.data 16494C-002

\B2211A\pcif\kelvin\4m.data 16494C-005

\B2211A\cable\triax\1_5m.data 16494A-001 16495F/G C2H, C2L, C3H, and C3L \B2211A\cable\triax\3m.data 16494A-002

\B2211A\cable\triax\4m.data 16494A-005

\B2211A\cable\kelvin\1_5m.data 16494B-001

\B2211A\cable\kelvin\3m.data 16494B-002

\B2211A\cable\kelvin\4m.data 16494C-005

a. <path>driver_install_folder\AGB220XA\ccdata (e.g. C:\temp\AGB220XA\ccdata)

b. Model number of the cable connected between the switch module and the DUT interface.

c. Agilent B2220A probe card interface or Agilent 16495F/G connector plate.

Show 5 more labels / short text on this page
Programming Capacitance Compensation · File namea · DUT interfacec · Coefficients to · be modified
Page 123 of the Agilent Technologies B2201A user manual — R ( Ω ) L (H) C (F)
Page 123 · click to enlarge
Page 123

Table 4-10 Compensation Coefficients and Modifications

coefficients Modifications of data file

C2H

For the Agilent B2220A probe card interface, do not modify the lines.

For the connector plate, change the R, L, C values in the lines. The value must be changed to the R, L, C values of the C2x path (triaxial cable with connector plate) shown in Figure 4-6.

C3H

Change the R, L, C values in the lines. The value must be changed to the R, L, C values of the C3x path.

For the Agilent B2220A probe card interface, probe card will be used for the C3x path.

For the connector plate, coaxial cable with positioner will be used for the C3x path.

2. Measure the R, L, C values of the C2x or C3x path by using the Agilent 4284A. See “To obtain compensation coefficients” on page 4-26.

After the measurements, calculate the per meter value of the R, L, C, and record them into the following table.

R ( Ω ) L (H) C (F)

C2H

C3H

3. Open the template file selected at step 1 by using a text editor. Exchange the R, L, C values of C2x/C3x with the values recorded at step 2. And save the file as your compensation data file (e.g. C:\temp\my_env_1.txt).

Do not change any other lines. Also do not change the value for the coefficients that should not be modified.

Show 5 more labels / short text on this page
Programming Capacitance Compensation · Compensation · Explanation · Compensation · coefficients
Page 124 of the Agilent Technologies B2201A user manual — To obtain compensation coefficients
Page 124 · click to enlarge
Page 124

To obtain compensation coefficients

Obtain the compensation coefficients as shown below.

1. Select the measurement frequency (Fmeas) used for the capacitance measurement of a device under test (DUT), and set it to the Agilent 4284A. The coefficients must be measured at the same frequency.
2. Perform the Agilent 4284A open calibration at the measurement terminal. Optionally, perform short calibration if you want.
3. See Table 4-11 and Figure 4-7, and set the Agilent 4284A.
4. Connect the path/cable corresponding to C3H shown in Figure 4-6 to the Agilent 4284A. Then measure and record the R, L, and C values.
5. Connect the path/cable corresponding to C3L to the Agilent 4284A. Then measure and record the R, L, and C values.
6. If you use the connector plate, perform the following procedure.

a. Connect the path/cable corresponding to C2H to the Agilent 4284A. Then measure and record the R, L, and C values.

b. Connect the path/cable corresponding to C2L to the Agilent 4284A. Then

measure and record the R, L, and C values.

Table 4-11 R, L, C Measurement Conditions

Agilent Technologies B2201A — To obtain compensation coefficients diagram (page 124)
Fig. 18 · To obtain compensation coefficients

a. Select 1 point. Do not change while measurements of all coefficients.

b. For triaxial cable, connect B to F directly, and measure L between A

and E. For coaxial cable, connect B to D directly, and measure L between A and C. Ignore E and F.

Figure 4-7 Measurement Terminals of C2H/C2L/C3H/C3L Path

Agilent Technologies B2201A — To obtain compensation coefficients diagram (page 124)
Fig. 19 · To obtain compensation coefficients

Programming Capacitance Compensation

Page 125 of the Agilent Technologies B2201A user manual — C (F) G (S)
Page 125 · click to enlarge
Page 125

To Perform Measurement and Compensation

Perform the capacitance measurement and compensation as shown below.

1. Set the Agilent 4284A measurement condition. Then the frequency must be the value (Fmeas) used when the compensation coefficients are measured.
2. Before contacting the device under test (DUT), perform the Cp-G measurement in the open condition at the end of the measurement path including positioner or probe card, and record the measurement data ( C1 and G1 ). See Table 4-12.
3. Perform the compensation by using the capacitance compensation function, and record the result data ( C1r and G1r ).

See Table 4-13 for the example to use the capacitance compensation function. This example uses Microsoft Visual Basic .NET and Agilent T&M Programmers Toolkit. And the Agilent B2200 controlled by the program is defined as AGB2200 at the 10th line. So, the function names are changed as shown below.

agb220xa_selectCompenFile --> Agb2200.SelectCompenFile

agb220xa_compenC --> Agb2200.CompenC

Also, agb220xa_reset and agb220xa_close have been changed.

4. Contact the DUT, perform the Cp-G measurement, and record the measurement data ( C2 and G2 ).
5. Perform the compensation and record the result data ( C2r and G2r ).
6. Perform the following calculation and record it as the capacitance value.

C = C2r - C1r

Table 4-12 Recording Measurement/Compensation Data

C (F) G (S)

2 (measured) C1 = G1 =

3 (compensated) C1r = G1r =

4 (measured) C2 = G2 =

5 (compensated) C2r = G2r =

6 (calculated) C =

Programming Capacitance Compensation · Measurement/Compensation Data · step

Page 126 of the Agilent Technologies B2201A user manual — To Perform Measurement and Compensation
Page 126 · click to enlarge
Page 126

Table 4-13 Capacitance Compensation Program Example

Imports Agilent.TMFramework Imports Agilent.TMFramework.DataAnalysis Imports Agilent.TMFramework.DataVisualization Imports Agilent.TMFramework.InstrumentIO Imports Agilent.TMFramework.InstrumentDriverInterop Imports Agilent.TMFramework.InstrumentDriverInterop.Design Imports Agilent.TMFramework.InstrumentDriverInterop.VxipnpWrappers Module Module1

Sub Main() ’9 Dim Agb2200 As Agb220xa = New Agb220xa("GPIB0::22::INSTR", True, True) Agb2200.Reset() Dim f_com as String = "C:\temp\my_env_1.txt" ’12 Agb2200.SelectCompenFile(f_com)

Dim freq As Double = 1000000 ’measurement frequency: 1 (MHz) 15 Dim data_c As Double = 0.0000000001 ’C measured by 4284A: 100 (pF) Dim data_g As Double = 0.0005 ’G measured by 4284A: 0.5 (mS) Dim res_c As Double Dim res_g As Double Agb2200.CompenC(freq, data_c, data_g, res_c, res_g) ’20

Dim result As String = "C = " & res_c * 1000000000000.0 & " (pF)" ’22 result = result & Chr(10) & "G = " & res_g * 1000 & " (mS)" & Chr(10) Console.WriteLine(result) Agb2200.Close() End Sub End Module

1 to 11 The above example is for the B2200 of the GPIB address 22 on the interface GPIB0. “GPIB0” is the VISA name. Confirm your GPIB settings, and set them properly.

12 to 13 The lines specify the compensation data file. The file name must specify your compensation data file.

15 to 20 Compensates the data measured by the Agilent 4284A. In this example, the measurement frequency is 1 MHz, the capacitance data is 100 pF, and the conductance data is 0.5 mS.

Change freq, data_c, data_g values for your measurement results.

22 to 24 Displays the compensation result data on the console window. Record the values as C1r and G1r, or C2r and G2r.

Programming Capacitance Compensation · Line Description

Page 127 of the Agilent Technologies B2201A user manual — 5 SCPI Command Reference
Page 127 · click to enlarge
Page 127

5 SCPI Command Reference

Page 128 of the Agilent Technologies B2201A user manual — 5 SCPI Command Reference
Page 128 · click to enlarge
Page 128

This chapter describes the following for Agilent B2200:

  • SCPI commands available to control the B2200 via GPIB interface.

SCPI is a universal programming language for electronic test and measurement instruments, and is based on IEEE 488.1 and IEEE 488.2.

SCPI commands are divided into two types: common commands and subsystem commands.

Common commands are generally not measurement related, but are used to manage status registers, data storage, and so on. Common commands are defined by IEEE 488.2. All common commands begin with an asterisk, such as *RST.

Subsystem commands are mostly measurement related and some are general purpose. Subsystem commands have a hierarchical structure that uses a colon between mnemonics (keywords), such as :DIAG:TEST:CARD:CLE.

Commands for the following subsystems are available for the B2200:

DIAGnostic commands for performing self-test.

ROUTe commands for signal routing.

SYSTem commands for controlling other functions that are not related to instrument performance.

Commands are listed in alphabetical order in this chapter.

  • Status reporting structure

These are IEEE 488.2 status structures (registers), which can be set or read by the common commands.

SCPI Command Reference

Page 129 of the Agilent Technologies B2201A user manual — 5 SCPI Command Reference
Page 129 · click to enlarge
Page 129

Textual Notation for Subsystem Commands

CAPITAL LETTERS Capital letters are the minimally required letters of the command header. Lowercase letters are the long form (complete spelling), which you can omit if desired.

For example, for :SYSTem:CCONfig?, you only need to specify :SYST:CCON?.

[ ] Square brackets are used to enclose optional information not required for execution of the command sequence.

For example, for the [:ROUTe]:BIAS:PORT command, you can execute :ROUT:BIAS:PORT or :BIAS:PORT, which are exactly the same.

italic font Text you supply. For example, for card_number, you enter an allowed card number. Also, used to represent response data.

| Vertical bar can be read as “or” and is used to separate alternative parameter options.

{ } Braces (curly brackets) are used to enclose one or more parameters that may be included zero or more times.

< > Angular brackets indicate that the word or words enclosed represent something other than themselves. For example, .

SCPI Command Reference

Page 130 of the Agilent Technologies B2201A user manual — Commands Summary
Page 130 · click to enlarge
Page 130

SCPI Command Reference Commands Summary

Common Commands

This section describes common commands and queries, which are commands defined by IEEE 488.2.

Commands Summary

The following table shows some common commands that are supported for the B2200.

*ESE(?) Standard Event Status Enable Command (Query)

*ESR? Standard Event Status Register Query

*OPC(?) Operation Complete Command (Query)

*SRE(?) Service Request Enable Command (Query)

*STB? Read Status Byte Query

Show 6 more labels / short text on this page
Mnemonic Name · *CLS Clear Status · *IDN? Identification Query · *RST Reset Command · *TST? Self-Test Query · *WAI Wait-to-Continue Command
Page 131 of the Agilent Technologies B2201A user manual — *ESE
Page 131 · click to enlarge
Page 131

*CLS

*CLS

This command clears the Status Byte Register, the Standard Event Status Register, and the Error Queue. This command does not clear the enable registers. See “Status Reporting Structure” on page 5-53.

Also, this command stops the monitoring of pending operations by the *OPC command.

This command does not have query form.

*ESE

This command sets the bits of the Standard Event Status “Enable” Register. 1 enables, 0 masks.

enable_number decimal integer (that is the sum of the binary-weighted values for the desired bits), hexadecimal, octal, or binary value

Query response enable_number

enable_number is decimal integer value that is the sum of the binary-weighted values of the “Enable” register bits.

Semantics The Standard Event Status “Enable” Register determines which bits of the Standard Event Status Register are enabled. Enabled bits are ORed together, and the result is reported to bit5 of the Status Byte Register.

The Standard Event Status “Enable” Register consists of 16 bits, but only the lower 8 bits are used, which correspond to the bits of the Standard Event Status Register. For details, see “Standard Event Status Enable Register” on page 5-59.

Show 5 more labels / short text on this page
SCPI Command Reference · Syntax *CLS · Example OUTPUT @Agb2200;"*CLS" · Syntax *ESE enable_number · Parameter Explanation
Page 132 of the Agilent Technologies B2201A user manual — *ESR?
Page 132 · click to enlarge
Page 132

SCPI Command Reference *ESR?

The following table shows the bits of the Standard Event Status Register and binary-weighted decimal value of each bit.

0 1 OPC (Operation Complete)

1 2 not used

2 4 QYE (Query ERROR)

3 8 DDE (Device-Dependent ERROR)

4 16 EXE (Execution ERROR)

5 32 CME (Command ERROR)

6 64 not used

7 128 PON (Power on)

Example The following four lines enable the same bit (CME bit):

OUTPUT @Agb2200;"*ESE 32" using decimal numeric

OUTPUT @Agb2200;"*ESE #B100000" using binary numeric

OUTPUT @Agb2200;"*ESE #Q40" using octal numeric

OUTPUT @Agb2200;"*ESE #H20" using hexadecimal numeric

The following is example for query:

OUTPUT @Agb2200;"*ESE?" ENTER @Agb2200;A

*ESR?

This query command returns the present contents of the Standard Event Status Register.

Query response register

register decimal integer value that is the sum of the binary-weighted values for the set bits

bit binary-weight description · Syntax *ESR? · Parameter Explanation

Page 133 of the Agilent Technologies B2201A user manual — AGILENT,B2200A,0,A.01.00
Page 133 · click to enlarge
Page 133

*IDN?

Semantics The following table shows the bits of the Standard Event Status Register, and the binary-weighted decimal value of each bit.

0 1 OPC (Operation Complete)

1 2 not used. always 0.

2 4 QYE (Query ERROR)

3 8 DDE (Device-Dependent ERROR)

4 16 EXE (Execution ERROR)

5 32 CME (Command ERROR)

6 64 not used. always 0.

7 128 PON (Power on)

Example OUTPUT @Agb2200;"*ESR?" ENTER @Agb2200;A

*IDN?

This query command returns the ID of your B2200.

Query response AGILENT,model,0,revision

model character model number. B2200A or B2201A

revision character revision number. A.01.00 or later

Example DIM A$[50] OUTPUT @Agb2200;"*IDN?" ENTER @Agb2200;A$ PRINT A$

An example result of above program is:

AGILENT,B2200A,0,A.01.00

Show 4 more labels / short text on this page
SCPI Command Reference · bit binary-weight description · Syntax *IDN? · Response Type Explanation
Page 134 of the Agilent Technologies B2201A user manual — *OPC
Page 134 · click to enlarge
Page 134

SCPI Command Reference *OPC

*OPC

This command starts to monitor pending operations, and sets/clears the Operation Complete (OPC) bit in the Standard Event Status Register as follows:

  • If there is no pending operation, sets the OPC bit to 1.
  • If there are any pending operations, sets the OPC bit to 0. The bit will be set to 1 again when all pending operations are completed.

So, *OPC command is required to enable the OPC bit. To stop monitoring pending operations (disable OPC bit), execute the *CLS command.

For details about the Standard Event Status Register, see “Standard Event Status Register” on page 5-58. Also, see *WAI command.

Query response 1

*OPC? places ASCII character 1 into the Output Queue when all pending operations are completed. For details, see “Output Queue” on page 5-60.

The following example is for query:

OUTPUT @Agb2200;"*OPC?" ENTER @Agb2200;A

Syntax *OPC · Example OUTPUT @Agb2200;"*OPC"

Page 135 of the Agilent Technologies B2201A user manual — *RST
Page 135 · click to enlarge
Page 135

*RST

*RST

This command performs an instrument reset. Status after *RST is shown below:

Channel ConfigurationAuto configuration mode
Connection RuleFree
Connection SequenceBreak Before Make
Bias ModeOff
Bias Input Port10

Bias-enabled Channels: All channels (output ports) are bias-enabled.

Couple ModeOff

Couple Input Port: Cleared.

Ground ModeOff
Ground Input Port12

Ground-enabled Channels: Cleared.

Ground-enabled Ports: Cleared.

Card Channel Status: All relays are opened.

Self-Test Result: Not changed.

Input Port Symbol String: Cleared.

Output Port Symbol String: Cleared.

BeeperOn
LCD in remote modeOff
LED in remote modeOn

Light Pen: Enabled.

For the power-on settings, see :SYSTem:CPON.

SCPI Command Reference · Syntax *RST · Example OUTPUT @Agb2200;"*RST"

Page 136 of the Agilent Technologies B2201A user manual — *SRE
Page 136 · click to enlarge
Page 136

SCPI Command Reference *SRE

*SRE

This command sets the Service Request “Enable” Register bits. 1 enables, 0 masks.

enable_number decimal integer (that is the sum of the binary-weighted values for the desired bits), hexadecimal, octal, or binary value

Query response enable_number

Semantics The Service Request “Enable” Register consists of 8 bits: Bit0 to Bit7. Bit6 is not defined, and is always 0. The Service Request “Enable” Register determines which bits of the Status Byte Register are enabled.

The status of the enabled bits are ORed together, and the result of OR is output to bit6 (Master Summary Status bit) of Status Byte Register. For details, see “Status Reporting Structure” on page 5-53.

The following table shows the bits of the Status Byte Register, and the binary-weighted decimal value of each bit.

0 1 not used

1 2 not used

2 4 not used

3 8 not used

4 16 MAV (Message Available summary-message)

5 32 ESB (Event Status Bit)

6 64 MSS (Master Summary Status)

7 128 not used

Example The following four lines enable the same bits (bit 4 and 5):

OUTPUT @Agb2200;"*SRE 48" using decimal numeric

OUTPUT @Agb2200;"*SRE #B110000" using binary numeric

OUTPUT @Agb2200;"*SRE #Q60" using octal numeric

OUTPUT @Agb2200;"*SRE #H30" using hexadecimal numeric

Syntax *SRE enable_number · Parameter Explanation · bit binary-weight description

Page 137 of the Agilent Technologies B2201A user manual — *STB?
Page 137 · click to enlarge
Page 137

*STB?

The following is example for query:

OUTPUT @Agb2200;"*SRE?" ENTER @Agb2200;A

*STB?

This query command reads the Status Byte Register (reads Master Summary Status bit, not Request for Service Message).

For bit6, this command reads MSS, not Request for Service (RQS). See “Status Reporting Structure” on page 5-53.

Query response register

register decimal integer value that is the sum of the binary-weighted values for the set bits

Semantics The following table shows the bits of the Status Byte Register, and the binary-weighted decimal value of each bit.

0 1 not used. always 0.

1 2 not used. always 0.

2 4 not used. always 0.

3 8 not used. always 0.

4 16 MAV (Message Available summary-message)

5 32 ESB (Event Status Bit)

6 64 MSS (Master Summary Status)

7 128 not used. always 0.

Example OUTPUT @Agb2200;"*STB?" ENTER @Agb2200;A

Show 4 more labels / short text on this page
SCPI Command Reference · Syntax *STB? · Parameter Explanation · bit binary-weight description
Page 138 of the Agilent Technologies B2201A user manual — *WAI
Page 138 · click to enlarge
Page 138

SCPI Command Reference *TST?

*TST?

This query command executes an internal self-test, then returns the result. After this command execution, the B2200 becomes same status as after *RST command execution.

Query response test_result

Example OUTPUT @Agb2200;"*TST?" ENTER @Agb2200;A

*WAI

This command stops execution of any commands until the Operation Complete (OPC) bit is set to 1, which means there is no pending operation. See *OPC command.

Show 6 more labels / short text on this page
Syntax *TST? · test_result Explanation · 0 pass · 1 fail · Syntax *WAI · Example OUTPUT @Agb2200;"*WAI"
Page 139 of the Agilent Technologies B2201A user manual — Command Summary
Page 139 · click to enlarge
Page 139

Subsystem Commands

Command Summary

ROUT subsystem:ROUTe subsystem has commands for controlling the signal routing.

“Open a channel”: opens relays to disconnect the channel (that is, disconnects input port from output port).

“Close a channel”: closes relays to connect the channel (that is, connects input port to output port).

[:ROUT]:FUNC channel_config [:ROUT]:FUNC?

Sets the channel configuration mode. channel_config: ACON: Auto Config Mode NCON: Normal Config Mode

Query returns the present channel configuration: ACON or NCON.

[:ROUT]:CONN:RULE card_number,rule [:ROUT]:CONN:RULE? card_number

Sets the connection rule for the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config rule: FREE (free) or SROUte (single)

Query returns the connection rule of the specified card: FREE or SROU. card_number: Card to check. 0 for Auto Config, 1, 2, 3, or 4 for Normal Config.

[:ROUT]:CONN:SEQ card_number,sequence [:ROUT]:CONN:SEQ? card_number

Specifies the connection sequence mode for the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config sequence: NSEQ: No-Sequence mode BBM: Break-Before-Make mode MBBR: Make-Before-Break mode

Query returns the connections sequence mode of the specified card: NSEQ, BBM, or MBBR. card_number: Card to check. 0 for Auto Config, 1, 2, 3, or 4 for Normal Config.

SCPI Command Reference · Command Summary · Command Description

Page 140 of the Agilent Technologies B2201A user manual — Subsystem Commands
Page 140 · click to enlarge
Page 140

SCPI Command Reference Command Summary

[:ROUT]:SYMB:CHAN card_number,channel,'string' [:ROUT]:SYMB:CHAN? card_number,channel

Defines a string for the specified channel.

card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config

channelchannel number, 1 to 48 for Auto, 1 to 12 for Normal

Query returns the symbol string set to the specified channel.

[:ROUT]:SYMB:PORT port,'string' [:ROUT]:SYMB:PORT? port

Defines a string for the specified input port.

portinput port number, 1 to 14

Query returns the symbol string set to the specified input port.

[:ROUT]:OPEN:CARD card_number Disconnects all input ports from all output ports for the specified card.

card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config

[:ROUT]:OPEN[:LIST] (@channel_list) [:ROUT]:OPEN[:LIST]? (@channel_list)

Disconnects the input ports from output ports as specified in channel_list.

channel_list: Channels to open.

Query returns the status of the specified channels: 0 (closed) or 1 (opened). channel_list: Channels to check.

[:ROUT]:CLOS:CARD? card_number Returns channel_list of all closed channels for the specified card. “closed channel” means an input port connected to an output port.

card_number: Card to check. 0 for Auto Config, 1, 2, 3, or 4 for Normal Config.

[:ROUT]:CLOS[:LIST] (@channel_list) [:ROUT]:CLOS[:LIST]? (@channel_list)

Connects the input ports to the output ports as specified in channel_list.

channel_list: Channels to close.

Query returns the status of the specified channels: 1 (closed) or 0 (opened). channel_list: Channels to check.

Command Description · Relay Control Commands

Page 141 of the Agilent Technologies B2201A user manual — Subsystem Commands
Page 141 · click to enlarge
Page 141

[:ROUT]:BIAS:CHAN:DIS:CARD card_number Bias-disables the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config

[:ROUT]:BIAS:CHAN:DIS[:LIST] (@channel_list) [:ROUT]:BIAS:CHAN:DIS[:LIST]? (@channel_list)

Bias-disables the specified channels. channel_list: Channels to bias-disable.

Query returns the status of the specified channels: 1 (disabled) or 0 (enabled). channel_list: Channels to check.

[:ROUT]:BIAS:CHAN:ENAB:CARD card_number Bias-enables the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config

[:ROUT]:BIAS:CHAN:ENAB[:LIST] (@channel_list) [:ROUT]:BIAS:CHAN:ENAB[:LIST]? (@channel_list)

Bias-enables the specified channels. channel_list: Channels to bias-enable.

Query returns the status of the specified channels: 1 (enabled) or 0 (disabled). channel_list: Channels to check.

[:ROUT]:BIAS:PORT card_number,bias_port [:ROUT]:BIAS:PORT? card_number

Specifies the input Bias Port for the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config bias_port: 1 to 14 or -1

Query returns the input Bias Port number for the specified card. card_number: Card to check. 0 for Auto Config, 1, 2, 3, or 4 for Normal Config.

[:ROUT]:BIAS[:STAT] card_number,state [:ROUT]:BIAS[:STAT]? card_number

Sets the bias mode for the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config state: ON / 1 (mode ON) or OFF / 0 (mode OFF)

Query returns the mode status of the specified card: 0 (OFF) or 1 (ON). card_number: Card to check. 0 for Auto Config, 1, 2, 3, or 4 for Normal Config.

Show 4 more labels / short text on this page
SCPI Command Reference · Command Summary · Command Description · Bias Mode Commands
Page 142 of the Agilent Technologies B2201A user manual — Subsystem Commands
Page 142 · click to enlarge
Page 142

SCPI Command Reference Command Summary

[:ROUT]:AGND:CHAN:DIS:CARD card_number Ground-disables the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config

[:ROUT]:AGND:CHAN:DIS[:LIST] (@channel_list) [:ROUT]:AGND:CHAN:DIS[:LIST]? (@channel_list)

Ground-disables the specified channels. channel_list: Channels to ground-disable.

Query returns the status of the specified channels: 1 (disabled) or 0 (enabled). channel_list: Channels to check.

[:ROUT]:AGND:CHAN:ENAB:CARD card_number Ground-enables the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config

[:ROUT]:AGND:CHAN:ENAB[:LIST] (@channel_list) [:ROUT]:AGND:CHAN:ENAB[:LIST]? (@channel_list)

Ground-enables the specified channels. channel_list: Channels to ground-enable.

Query returns the status of the specified channels: 1 (enabled) or 0 (disabled). channel_list: Channels to check.

[:ROUT]:AGND:PORT card_number,ground_port [:ROUT]:AGND:PORT? card_number

Specifies the input Ground Port for the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config ground_port: 1 to 14 or -1

Query returns the input Ground Port number for the specified card. card_number: Card to check. 0 for Auto Config, 1, 2, 3, or 4 for Normal Config.

[:ROUT]:AGND[:STAT] card_number,state [:ROUT]:AGND[:STAT]? card_number

Sets the ground mode for the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config state: ON / 1 (mode ON) or OFF / 0 (mode OFF)

Query returns the mode status of the specified card: 0 (OFF) or 1 (ON). card_number: Card to check. 0 for Auto Config, 1, 2, 3, or 4 for Normal Config.

Command Description · Ground Mode Commands

Page 143 of the Agilent Technologies B2201A user manual — Subsystem Commands
Page 143 · click to enlarge
Page 143

[:ROUT]:AGND:UNUSED card_number,'enable_port' [:ROUT]:AGND:UNUSED? card_number

Ground-enables the specified input ports for the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config enable_port: One or more input port numbers: 1 to 8. Enclose by single quotation marks. Separate multiple input port numbers by comma. For example: '1,5'

Query returns the ground-enabled input port numbers for the specified card. card_number: Card to check. 0 for Auto Config, 1, 2, 3, or 4 for Normal Config.

[:ROUT]:COUP:PORT card_number,'couple_port' [:ROUT]:COUP:PORT? card_number

Specifies the input couple ports for the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config couple_port: One or more input port numbers: 1, 3, 5, 7, 9, 11, or 13. Enclose by single quotation marks. Separate multiple input port numbers by comma. For example: '1,5'

Query returns the lower input port number of each couple pair for the specified card. card_number: Card to check. 0 for Auto Config, 1, 2, 3, or 4 for Normal Config.

[:ROUT]:COUP:PORT:DET Detects the input ports connected to Kelvin cable, and assigns them as the input couple ports for the all cards.

[:ROUT]:COUP[:STAT] card_number,state [:ROUT]:COUP[:STAT]? card_number

Sets the couple mode for the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config state: ON / 1 (mode ON) or OFF / 0 (mode OFF)

Query returns the mode status of the specified card: 0 (OFF) or 1 (ON). card_number: Card to check. 0 for Auto Config, 1, 2, 3, or 4 for Normal Config.

Show 4 more labels / short text on this page
SCPI Command Reference · Command Summary · Command Description · Couple Mode Commands
Page 144 of the Agilent Technologies B2201A user manual — Subsystem Commands
Page 144 · click to enlarge
Page 144

SCPI Command Reference Command Summary

DIAG subsystem:DIAGnostic subsystem has commands for executing the self-test function. For more info, see “Selftest Menu” on page 3-34.

The :DIAGnostic subsystem commands ignore the B2200 channel configuration mode. For :DIAG commands that require a card number, you specify 1, 2, 3, 4, or ALL.

:DIAG:TEST:CARD:CLE card_number Clears relay test result (pass/fail) of the specified card.

card_number1 , 2 , 3 , 4 , or ALL

:DIAG:TEST:CARD[:EXEC]? card_number Executes relay test, then returns result: 1 (fail card exists), 0 (pass).

card_number1 , 2 , 3 , 4 , or ALL

:DIAG:TEST:CARD:STAT? card_number Returns most recent relay test result: 1 (fail), 0 (pass), -1 (not tested).

card_number1 , 2 , 3 , 4

:DIAG:TEST:FRAM:CLE item Clears specified test result.

itemCONT (controller test), FPAN (front panel interface test), LED, PEN, or BEEP

:DIAG:TEST:FRAM[:EXEC]? item Executes specified test, then returns test result: 1 (fail), 0 (pass).

itemCONT (controller test), FPAN (front panel interface test), LED, PEN, or BEEP

:DIAG:TEST:FRAM:STAT? item Returns most recent test result of the specified test: 1 (fail), 0 (pass), -1 (not tested).

itemCONT (controller test), FPAN (front panel interface test), LED, PEN, or BEEP

Command Description

Page 145 of the Agilent Technologies B2201A user manual — Subsystem Commands
Page 145 · click to enlarge
Page 145

:SYSTem subsystem is a collection of functions that are not related to instrument performance.

:SYST:BEEP state Enables/disables the beeper. state: ON / 1 (enable) or OFF / 0 (disable)

:SYST:CCON? card_number Returns the card configuration information. This command is just to keep compatibility with the Agilent E5250A. card_number: 1, 2, 3, or 4

:SYST:CDES? card_number Returns a description of the specified card: model number and input/output port information. card_number: Card to check. 0 for Auto Config, 1, 2, 3, or 4 for Normal Config.

:SYST:CPON card_number Resets the specified card to the power-on state. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config

:SYST:CTYP? card_number Returns ID of the specified card: model number and revision. card_number: Card to check. 0 for Auto Config, 1, 2, 3, or 4 for Normal Config.

:SYST:DISP:LCD state Enables/disables the front panel LCD when the B2200 is in the GPIB remote mode. state: ON / 1 (enable) or OFF / 0 (disable)

:SYST:DISP:LED state Enables/disables the front panel LED. state: ON / 1 (enable) or OFF / 0 (disable)

:SYST:DISP:STR string Specifies a string displayed on the LCD in the GPIB remote mode.

:SYST:ERR? Reads error from head of error queue, and removes it from the queue.

:SYST:KLC state Locks/unlocks the front panel keys. state: ON / 1 (lock) or OFF / 0 (unlock)

:SYST:MEMO:SAVE memory_number:SYST:MEMO:LOAD memory_number

Saves a setup information into the internal memory, or loads a setup information. memory_number: 1 to 8

Show 4 more labels / short text on this page
SCPI Command Reference · Command Summary · SYSTEM subsystem · Command Description
Page 146 of the Agilent Technologies B2201A user manual — Subsystem Commands
Page 146 · click to enlarge
Page 146

SCPI Command Reference Command Summary

:SYST:MEMO:COMM memory_number,'comment' :SYST:MEMO:COMM? memory_number

Memorizes the comment for the B2200 setup information specified by memory_number. memory_number: 1 to 8

:SYST:MEMO:DEL memory_number Deletes the B2200 setup information and the comment specified by memory_number. memory_number: 1 to 8

:SYST:PEN state Enables/disables the light pen. state: ON / 1 (enable) or OFF / 0 (disable)

:SYST:VERS? Returns SCPI version number for which the B2200 complies.

Command Description

Page 147 of the Agilent Technologies B2201A user manual — :DIAGnostic:TEST:CARD[:EXECute]?
Page 147 · click to enlarge
Page 147

SCPI Command Reference :DIAGnostic:TEST:CARD:CLEar

:DIAGnostic:TEST:CARD:CLEar

This command clears the relay test result (pass/fail result) of the specified card.

Syntax:DIAGnostic:TEST:CARD:CLEar card_number

card_number card number1 , 2 , 3 , 4 , or ALL

Example OUTPUT @Agb2200;":DIAG:TEST:CARD:CLE 1"

:DIAGnostic:TEST:CARD[:EXECute]?

This command executes the relay test for specified card, then returns the pass/fail result. Before starting the relay test, open the input/output terminals (end of cable is OK, Kelvin cable must be removed). For more information, see “RELAY_TEST” on page 3-35.

Syntax :DIAGnostic:TEST:CARD[:EXECute]? card_number

card_number card number1 , 2 , 3 , 4 , or ALL

Query response test_result

1. failure card exists 0 : pass

Example OUTPUT @Agb2200;":DIAG:TEST:CARD? ALL"

Parameter Explanation · Parameter Explanation

Page 148 of the Agilent Technologies B2201A user manual — :DIAGnostic:TEST:FRAMe:CLEar
Page 148 · click to enlarge
Page 148

SCPI Command Reference :DIAGnostic:TEST:CARD:STATe?

:DIAGnostic:TEST:CARD:STATe?

This command returns the most recent relay test result for the specified card.

Syntax :DIAGnostic:TEST:CARD:STATe? card_number

card_number card number1 , 2 , 3 , 4

Query response test_result

1. fail 0 : pass -1 : not tested

Example OUTPUT @Agb2200;":DIAG:TEST:CARD:STAT? 1" ENTER @Agb2200;A

:DIAGnostic:TEST:FRAMe:CLEar

This command clears test result of the specified B2200 test.

Syntax :DIAGnostic:TEST:FRAMe:CLEar CONTroller | FPANel | LED | PEN | BEEPer

FPANel Front panel interface test

LED LED matrix test

PEN Light pen test

Example OUTPUT @Agb2200;":DIAG:TEST:FRAM:CLE CONT"

Show 4 more labels / short text on this page
Parameter Explanation · Parameter Explanation · CONTroller Controller test · BEEPer Beeper test
Page 149 of the Agilent Technologies B2201A user manual — :DIAGnostic:TEST:FRAMe[:EXECute]?
Page 149 · click to enlarge
Page 149

SCPI Command Reference :DIAGnostic:TEST:FRAMe[:EXECute]?

:DIAGnostic:TEST:FRAMe[:EXECute]?

This command executes the specified B2200 test, then returns the test result. See “Selftest Menu” on page 3-34 to perform the test.

After the controller test, the B2200 status becomes same as after *RST command execution.

Syntax :DIAGnostic:TEST:FRAMe[:EXECute]? CONTroller | FPANel | LED | PEN | BEEPer

FPANel Front panel interface test

LED LED matrix test

PEN Light pen test

Query response test_result

1. fail 0 : pass

Example OUTPUT @Agb2200;":DIAG:TEST:FRAM? CONT"

Parameter Explanation · CONTroller Controller test · BEEPer Beeper test

Page 150 of the Agilent Technologies B2201A user manual — [:ROUTe]:AGND:CHANnel:DISable:CARD
Page 150 · click to enlarge
Page 150

SCPI Command Reference :DIAGnostic:TEST:FRAMe:STATe?

:DIAGnostic:TEST:FRAMe:STATe?

This command returns the most recent test result (pass/fail) of the specified test.

Syntax :DIAGnostic:TEST:FRAMe:STATe? CONTroller | FPANel | LED | PEN | BEEPer

FPANel Front panel interface test

LED LED matrix test

PEN Light pen test

Query response test_result

1. fail 0 : pass -1 : not tested

Example OUTPUT @Agb2200;":DIAG:TEST:FRAM:STAT? CONT" ENTER @Agb2200;A

[:ROUTe]:AGND:CHANnel:DISable:CARD

This command ground-disables the all output ports (channels) for the specified card. When the Ground Mode is ON, the ground-disabled output ports are disconnected from the input Ground Port. At *RST, no channel is ground-enabled. The ground mode is set by “[:ROUTe]:AGND[:STATe]”.

Syntax [:ROUTe]:AGND:CHANnel:DISable:CARD card_number

card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL

Example OUTPUT @Agb2200;":ROUT:AGND:CHAN:DIS:CARD ALL"

Show 4 more labels / short text on this page
Parameter Explanation · CONTroller Controller test · BEEPer Beeper test · Parameter Explanation
Page 151 of the Agilent Technologies B2201A user manual — [:ROUTe]:AGND:CHANnel:ENABle:CARD
Page 151 · click to enlarge
Page 151

SCPI Command Reference [:ROUTe]:AGND:CHANnel:DISable[:LIST]

[:ROUTe]:AGND:CHANnel:DISable[:LIST]

This command specifies the ground-disabled output ports (channels). When the Ground Mode is ON, the ground-disabled output ports are disconnected from the input Ground Port. At *RST, no channel is ground-enabled. The ground mode is set by “[:ROUTe]:AGND[:STATe]”.

The query returns whether the specified channels are ground-disabled or not.

Syntax [:ROUTe]:AGND:CHANnel:DISable[:LIST] (@channel_list)

[:ROUTe]:AGND:CHANnel:DISable[:LIST]? (@channel_list)

channel_list Channels to ground-disable or to check. For channel_list, see “Switch Control” on page 4-5.

Input port is always the input Ground Port. So, the input ports in channel_list are ignored. However, you cannot abbreviate the input port.

Query response disable_status{, disable_status}

1. ground disabled 0 : ground enabled

Example OUTPUT @Agb2200;":ROUT:AGND:CHAN:DIS (@10101)" OUTPUT @Agb2200;":ROUT:AGND:CHAN:DIS? (@10101,10102,10201)" ENTER @Agb2200;A$

This example ground-disables the output port 1 on the card 1. In this example, A$ will be 1,0,1.

[:ROUTe]:AGND:CHANnel:ENABle:CARD

This command ground-enables the all output ports (channels) for the specified card. When the Ground Mode is ON, the ground-enabled output ports that have not been connected to any other input port are connected to the input Ground Port. This command is available only for the specified card. At *RST, no channel is ground-enabled. The ground mode is set by “[:ROUTe]:AGND[:STATe]”.

Parameter Explanation

Page 152 of the Agilent Technologies B2201A user manual — [:ROUTe]:AGND:CHANnel:ENABle[:LIST]
Page 152 · click to enlarge
Page 152

SCPI Command Reference [:ROUTe]:AGND:CHANnel:ENABle[:LIST]

Syntax [:ROUTe]:AGND:CHANnel:ENABle:CARD card_number

card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL

Example OUTPUT @Agb2200;":ROUT:AGND:CHAN:ENAB:CARD ALL"

[:ROUTe]:AGND:CHANnel:ENABle[:LIST]

This command specifies the ground-enabled output ports (channels). When the Ground Mode is ON, the ground-enabled output ports that have not been connected to any other input port are connected to the input Ground Port. Then, the input Ground Port will not be connected to the output ports that have been connected to any other input port. At *RST, no channel is ground-enabled. The ground mode is set by “[:ROUTe]:AGND[:STATe]”.

The query returns whether the specified channels are ground-enabled or not.

Syntax [:ROUTe]:AGND:CHANnel:ENABle[:LIST] (@channel_list)

[:ROUTe]:AGND:CHANnel:ENABle[:LIST]? (@channel_list)

channel_list Channels to ground-enable or to check. For channel_list, see “Switch Control” on page 4-5.

Input port is always the input Ground Port. So, the input ports in channel_list are ignored. However, you cannot abbreviate the input port number.

Query response enable_status{, enable_status}

1. ground enabled 0 : ground disabled

Example OUTPUT @Agb2200;":ROUT:AGND:CHAN:ENAB (@10101)" OUTPUT @Agb2200;":ROUT:AGND:CHAN:ENAB? (@10101,10102,10201)" ENTER @Agb2200;A$

This example ground-enables the output port 1 on the card 1. In this example, A$ will be 1,0,1.

Parameter Explanation · Parameter Explanation

Page 153 of the Agilent Technologies B2201A user manual — [:ROUTe]:AGND:PORT
Page 153 · click to enlarge
Page 153

[:ROUTe]:AGND:PORT

[:ROUTe]:AGND:PORT

This command specifies the input Ground Port for the specified card. For each card, you can specify the same or different Ground Port. At *RST, the Ground Port is 12. The ground mode is set by “[:ROUTe]:AGND[:STATe]”.

The query returns the input port number of the Ground Port.

If the input ground port and a couple port have been assigned to the same input port, the ground mode and the couple mode cannot be used in parallel.

0 V output voltage source should be connected to the input Ground Port. Or set ground_port to 12, and open the input port 12 that is the default ground port. If you assign the input 12 as the ground port, this ground port will be internally connected to the ground when the ground mode is set to ON.

Syntax [:ROUTe]:AGND:PORT card_number,ground_port

[:ROUTe]:AGND:PORT? card_number

card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL ALL is not available for query.

ground_port Input port number: 1 to 14. Or -1 to disable ground port.

Query response port_number

port_number = 1 to 14 or -1. Number set to ground_port.

Example OUTPUT @Agb2200;":ROUT:AGND:PORT 0,13" OUTPUT @Agb2200;":ROUT:AGND:PORT? 0" ENTER @Agb2200;A

In this example, A will be 13.

SCPI Command Reference · Parameter Explanation

Page 154 of the Agilent Technologies B2201A user manual — [:ROUTe]:AGND[:STATe]
Page 154 · click to enlarge
Page 154

SCPI Command Reference [:ROUTe]:AGND[:STATe]

[:ROUTe]:AGND[:STATe]

This command controls the Ground Mode for the specified card. When the Ground Mode is ON, the input Ground Port is connected to the all ground-enabled input ports/output ports that have not been connected to any other port. At *RST, the Ground Mode is OFF.

The query returns the mode status.

The input Ground Port is set by “[:ROUTe]:AGND:PORT”. The ground-enabled input ports are set by “[:ROUTe]:AGND:UNUSED”. And the ground-enabled output ports are set by “[:ROUTe]:AGND:CHANnel:ENABle[:LIST]” or “[:ROUTe]:AGND:CHANnel:ENABle:CARD”.

When the Ground Mode is ON, you cannot directly control the Ground Port connections. However, they can be indirectly controlled when the connections of the other input port are controlled.

When the Ground Mode is OFF, the input Ground Port is the same as the other input ports, so relays can be controlled directly to connect to output ports.

Syntax [:ROUTe]:AGND[:STATe] card_number,state

[:ROUTe]:AGND[:STATe]? card_number

card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL ALL is not available for query.

state ON or 1sets the Ground Mode ON OFF or 0 : sets the Ground Mode OFF

Query response state

0. OFF or 1 : ON

Example OUTPUT @Agb2200;":ROUT:AGND:STAT 0,ON" OUTPUT @Agb2200;":ROUT:AGND:STAT? 0" ENTER @Agb2200;A

In this example, A will be 1.

Parameter Explanation

Page 155 of the Agilent Technologies B2201A user manual — [:ROUTe]:AGND:UNUSED
Page 155 · click to enlarge
Page 155

[:ROUTe]:AGND:UNUSED

[:ROUTe]:AGND:UNUSED

This command specifies the ground-enabled input ports for the specified card. When the Ground Mode is ON, the ground-enabled input ports that have not been connected to any other port are connected to the input Ground Port. At *RST, no input port is ground-enabled. The ground mode is set by “[:ROUTe]:AGND[:STATe]”.

The query returns which input ports are ground-enabled.

If the ground enabled input port and a couple port have been assigned to the same input port, the ground mode and the couple mode cannot be used in parallel.

Syntax [:ROUTe]:AGND:UNUSED card_number,'port_number'

[:ROUTe]:AGND:UNUSED? card_number

card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL ALL is not available for query.

port_number Input port number: 1 to 8.

Enclose by single quotation marks. Separate multiple input port numbers by comma. For example: '1,5'

Query response port_number{, port_number}

port_number = 1 to 8

Example OUTPUT @Agb2200;":ROUT:AGND:UNUSED 0,'5,6,7,8'" OUTPUT @Agb2200;":ROUT:AGND:UNUSED? 0" ENTER @Agb2200;A$

In this example, A$ will be 5,6,7,8.

SCPI Command Reference · Parameter Explanation

Page 156 of the Agilent Technologies B2201A user manual — [:ROUTe]:BIAS:CHANnel:DISable[:LIST]
Page 156 · click to enlarge
Page 156

SCPI Command Reference [:ROUTe]:BIAS:CHANnel:DISable:CARD

[:ROUTe]:BIAS:CHANnel:DISable:CARD

This command bias-disables the all output ports (channels) for the specified card. When the Bias Mode is ON, the bias-disabled output ports are disconnected from the input Bias Port. At *RST, all cards are bias-enabled. The bias mode is set by “[:ROUTe]:BIAS[:STATe]”.

Syntax [:ROUTe]:BIAS:CHANnel:DISable:CARD card_number

card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL

Example OUTPUT @Agb2200;":ROUT:BIAS:CHAN:DIS:CARD ALL"

[:ROUTe]:BIAS:CHANnel:DISable[:LIST]

This command specifies the bias-disabled output ports (channels). When the Bias Mode is ON, the bias-disabled output ports are disconnected from the input Bias Port. At *RST, all channels are bias-enabled. The bias mode is set by “[:ROUTe]:BIAS[:STATe]”.

The query returns whether the specified channels are bias-disabled or not.

Syntax [:ROUTe]:BIAS:CHANnel:DISable[:LIST] (@channel_list)

[:ROUTe]:BIAS:CHANnel:DISable[:LIST]? (@channel_list)

channel_list Channels to bias-disable or to check. For channel_list, see “Switch Control” on page 4-5.

Input port is always the input Bias Port. So, the input ports in channel_list are ignored. However, you cannot abbreviate the input port.

Query response disable_status{, disable_status}

1. bias disabled 0 : bias enabled

Parameter Explanation · Parameter Explanation

Page 157 of the Agilent Technologies B2201A user manual — [:ROUTe]:BIAS:CHANnel:ENABle[:LIST]
Page 157 · click to enlarge
Page 157

SCPI Command Reference [:ROUTe]:BIAS:CHANnel:ENABle:CARD

Example OUTPUT @Agb2200;":ROUT:BIAS:CHAN:ENAB:CARD ALL" OUTPUT @Agb2200;":ROUT:BIAS:CHAN:DIS (@10101)" OUTPUT @Agb2200;":ROUT:BIAS:CHAN:DIS? (@10101,10102,10201)" ENTER @Agb2200;A$

This example bias-disables the output port 1 on the card 1. In this example, A$ will be 1,0,1.

[:ROUTe]:BIAS:CHANnel:ENABle:CARD

This command bias-enables the all output ports (channels) for the specified card. When the Bias Mode is ON, the bias-enabled output ports that have not been connected to any other input port are connected to the input Bias Port. This command is available only for the specified card. At *RST, all cards are bias-enabled. The bias mode is set by “[:ROUTe]:BIAS[:STATe]”.

Syntax [:ROUTe]:BIAS:CHANnel:ENABle:CARD card_number

card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL

Example OUTPUT @Agb2200;":ROUT:BIAS:CHAN:ENAB:CARD ALL"

[:ROUTe]:BIAS:CHANnel:ENABle[:LIST]

This command specifies the bias-enabled output ports (channels). When the Bias Mode is ON, the bias-enabled output ports that have not been connected to any other input port are connected to the input Bias Port. Then, the input Bias Port will not be connected to the output ports that have been connected to any other input port. At *RST, all channels are bias-enabled. The bias mode is set by “[:ROUTe]:BIAS[:STATe]”.

The query returns whether the specified channels are bias-enabled or not.

Syntax [:ROUTe]:BIAS:CHANnel:ENABle[:LIST] (@channel_list)

[:ROUTe]:BIAS:CHANnel:ENABle[:LIST]? (@channel_list)

channel_list Channels to bias-enable or to check. For channel_list, see “Switch Control” on page 4-5.

Parameter Explanation · Parameter Explanation

Page 158 of the Agilent Technologies B2201A user manual — [:ROUTe]:BIAS:PORT
Page 158 · click to enlarge
Page 158

SCPI Command Reference [:ROUTe]:BIAS:PORT

Input port is always the input Bias Port. So, the input ports in channel_list are ignored. However, you cannot abbreviate the input port.

Query response enable_status{, enable_status}

1. bias enabled 0 : bias disabled

Example OUTPUT @Agb2200;":ROUT:BIAS:CHAN:DIS:CARD ALL" OUTPUT @Agb2200;":ROUT:BIAS:CHAN:ENAB (@10101)" OUTPUT @Agb2200;":ROUT:BIAS:CHAN:ENAB? (@10101,10102,10201)" ENTER @Agb2200;A$

This example bias enables the output port 1 on the card 1. In this example, A$ will be 1,0,1.

[:ROUTe]:BIAS:PORT

This command specifies the input Bias Port for the specified card. For each card, you can specify the same or different Bias Port. At *RST, Bias Port is set to 10. The bias mode is set by “[:ROUTe]:BIAS[:STATe]”. The query returns the input Bias Port number for the specified card.

Syntax [:ROUTe]:BIAS:PORT card_number,bias_port

[:ROUTe]:BIAS:PORT? card_number

card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL ALL is not available for query.

bias_port Input port number: 1 to 14. Or -1 to disable bias port.

Query response port_number

port_number = 1 to 14 or -1. Number set to bias_port.

Example OUTPUT @Agb2200;":ROUT:BIAS:PORT ALL,4" OUTPUT @Agb2200;":ROUT:BIAS:PORT? 1" ENTER @Agb2200;A

Parameter Explanation

Page 159 of the Agilent Technologies B2201A user manual — 0 : OFF 1 : ON
Page 159 · click to enlarge
Page 159

[:ROUTe]:BIAS[:STATe]

In this example, A will be 4.

[:ROUTe]:BIAS[:STATe]

This command controls the Bias Mode for the specified card. When the Bias Mode is ON, the input Bias Port is connected to the all bias-enabled output ports that have not been connected to any other input port. At *RST, the Bias Mode is OFF.

The query returns the mode status of the specified card.

The input Bias Port is set by “[:ROUTe]:BIAS:PORT”. The bias-enabled output ports (channels) are set by “[:ROUTe]:BIAS:CHANnel:ENABle[:LIST]” or “[:ROUTe]:BIAS:CHANnel:ENABle:CARD”.

When the Bias Mode is ON, you cannot directly control the Bias Port connections. However, they can be indirectly controlled when the connections of the other input port are controlled.

When the Bias Mode is OFF, the input Bias Port is the same as the other input ports, so relays can be controlled directly to connect to output ports.

Syntax [:ROUTe]:BIAS[:STATe] card_number,state

[:ROUTe]:BIAS[:STATe]? card_number

card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL ALL is not available for query.

state ON or 1sets the Bias Mode ON OFF or 0 : sets the Bias Mode OFF

Query response 0 or 1

0 : OFF 1 : ON

Example OUTPUT @Agb2200;":ROUT:BIAS:STAT ALL,ON" OUTPUT @Agb2200;":ROUT:BIAS:STAT? 4" ENTER @Agb2200;A

In this example, A will be 1.

SCPI Command Reference · Parameter Explanation

Page 160 of the Agilent Technologies B2201A user manual — [:ROUTe]:CLOSe[:LIST]
Page 160 · click to enlarge
Page 160

SCPI Command Reference [:ROUTe]:CLOSe:CARD?

[:ROUTe]:CLOSe:CARD?

This query command returns channel_list of all closed (connected) channels for the specified card.

Syntax [:ROUTe]:CLOSe:CARD? card_number

card_number For Auto Config mode0 For Normal Config mode: 1 , 2 , 3 , or 4

Query response channel_list

Example OUTPUT @Agb2200;":ROUT:OPEN:CARD ALL" OUTPUT @Agb2200;":ROUT:CLOS (@10101,10202)" OUTPUT @Agb2200;":ROUT:CLOS:CARD? 1" ENTER @Agb2200;A$

A$ will be @10101,10202. For channel_list, see “Switch Control” on page 4-5.

[:ROUTe]:CLOSe[:LIST]

This command connects the input ports to the output ports (channels) as specified in channel_list. The query returns 0 or 1 for each channel specified by channel_list, in same order as specified by channel_list.

For some modes (such as Single Route Mode, Couple Mode, Ground Mode, and Bias Mode), more complex connections may occur.

Syntax [:ROUTe]:CLOSe[:LIST] (@channel_list)

[:ROUTe]:CLOSe[:LIST]? (@channel_list)

channel_list Channels to close or to check. For channel_list, see “Switch Control” on page 4-5.

Query response close_status{, close_status}

1. closed 0 : opened

Parameter Explanation · Parameter Explanation

Page 161 of the Agilent Technologies B2201A user manual — [:ROUTe]:CONNection:RULE
Page 161 · click to enlarge
Page 161

SCPI Command Reference [:ROUTe]:CONNection:RULE

Example OUTPUT @Agb2200;":ROUT:CLOS (@10101,10202)" OUTPUT @Agb2200;":ROUT:CLOS? (@10101,10102,10201,10202)" ENTER @Agb2200;A$

In this example, A$ will be 1,0,0,1.

[:ROUTe]:CONNection:RULE

This command sets the connection rule (Free or Single Route) for the specified card. The query returns the connection rule of the specified card. At *RST, all cards are set to FREE. See “Connection Rule” on page 3-14.

Syntax [:ROUTe]:CONNection:RULE card_number,rule

[:ROUTe]:CONNection:RULE? card_number

card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL ALL is not available for query.

rule FREE (free route) or SROUte (single route).

In the free route, multiple channel connections are available for all input ports.

In the single route, only 1 channel connection is available for each input port.

Query response FREE or SROU

Example OUTPUT @Agb2200;":ROUT:CONN:RULE ALL,SROU" OUTPUT @Agb2200;":ROUT:CONN:RULE? 1" ENTER @Agb2200;A$

In this example, A$ will be SROU.

Parameter Explanation

Page 162 of the Agilent Technologies B2201A user manual — [:ROUTe]:CONNection:SEQuence
Page 162 · click to enlarge
Page 162

SCPI Command Reference [:ROUTe]:CONNection:SEQuence

[:ROUTe]:CONNection:SEQuence

This command specifies the connection sequence mode, which is the open/close sequence of the relays when connection route is changed from an existing connection to a new connection. The query returns the connection sequence mode of the specified card. At *RST, BBMake is selected. See “Connection Sequence” on page 3-15.

The connection sequence is used only for cards that have been set to SROUte connection rule. Also see “[:ROUTe]:CONNection:RULE”.

Syntax [:ROUTe]:CONNection:SEQuence card_number,sequence

[:ROUTe]:CONNection:SEQuence? card_number

card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL ALL is not available for query.

sequence NSEQNo-Sequence mode BBM : Break-Before-Make mode MBBR : Make-Before-Break mode

Query response NSEQ or BBM or MBBR

Example OUTPUT @Agb2200;":ROUT:CONN:SEQ ALL,MBBR" OUTPUT @Agb2200;":ROUT:CONN:SEQ? 0" ENTER @Agb2200;A$

In this example, A$ will be MBBR.

Parameter Explanation

Page 163 of the Agilent Technologies B2201A user manual — [:ROUTe]:COUPle:PORT
Page 163 · click to enlarge
Page 163

[:ROUTe]:COUPle:PORT

[:ROUTe]:COUPle:PORT

If the input ground port and a couple port have been assigned to the same input port, the ground mode and the couple mode cannot be used in parallel.

If the ground enabled input port and a couple port have been assigned to the same input port, the ground mode and the couple mode cannot be used in parallel.

This command is used to set up input couple ports, which are used for the Kelvin connections. The couple mode is set by “[:ROUTe]:COUPle[:STATe]”.

For each card, you can set up the same or different input couple ports.

This command overwrites the previous couple port setting. The couple port setting will be cleared and updated by “[:ROUTe]:COUPle:PORT:DETect”.

The query returns the odd input port number of each coupled pair.

At *RST, no input ports are coupled.

Syntax [:ROUTe]:COUPle:PORT card_number,'couple_port'

[:ROUTe]:COUPle:PORT? card_number

card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL ALL is not available for query.

couple_port One or more input port numbers: 1, 3, 5, 7, 9, 11 or 13.

Enclose by single quotation marks. Separate multiple input port numbers by comma. For example: '1,5'

Query response port_number{, port_number}

port_number = 1 or 3 or 5 or 7 or 9 or 11 or 13

Example OUTPUT @Agb2200;":ROUT:COUP:PORT ALL,'1,3'" OUTPUT @Agb2200;":ROUT:COUP:PORT? 1" ENTER @Agb2200;A$

In this example, A$ will be 1,3.

SCPI Command Reference · Parameter Explanation

Page 164 of the Agilent Technologies B2201A user manual — [:ROUTe]:COUPle:PORT:DETect
Page 164 · click to enlarge
Page 164

SCPI Command Reference [:ROUTe]:COUPle:PORT:DETect

[:ROUTe]:COUPle:PORT:DETect

This command detects the input ports connected to the Kelvin cable, and assigns them as the input couple ports that will be used for the Kelvin connection. The input couple port setting is effective for the all cards. This command overwrites the previous couple port setting.

The couple mode is set by “[:ROUTe]:COUPle[:STATe]”. The couple port setting will be cleared and updated by “[:ROUTe]:COUPle:PORT”.

The input port numbers of the couple ports can be got by [:ROUTe]:COUPle:PORT?.

After this command is executed, the all relay settings will be the same as after the :ROUT:OPEN:CARD ALL command execution.

Syntax [:ROUTe]:COUPle:PORT:DETect

Example OUTPUT @Agb2200;":ROUT:COUP:PORT:DET"

Page 165 of the Agilent Technologies B2201A user manual — 0 : OFF 1 : ON
Page 165 · click to enlarge
Page 165

[:ROUTe]:COUPle[:STATe]

[:ROUTe]:COUPle[:STATe]

This command controls the Couple Mode for the specified card. The query returns the mode status of the specified card. At *RST, the Couple Mode is OFF.

The input couple ports are set by “[:ROUTe]:COUPle:PORT:DETect” or “[:ROUTe]:COUPle:PORT”.

If you specify a connection from a couple input port to an output port when the couple mode is ON, the B2200 automatically controls relays to connect the input ports n-1 and n to the output ports m-1 and m respectively (n: an even number from 2 to 14, m: an even number from 2 to 12, 2 to 24, 2 to 36, or 2 to 48 depends on the B2200 configuration). See example below.

Syntax [:ROUTe]:COUPle[:STATe] card_number,state

[:ROUTe]:COUPle[:STATe]? card_number

card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL ALL is not available for query.

state ON or 1sets the Couple Mode ON OFF or 0 : sets the Couple Mode OFF

Query response state

0 : OFF 1 : ON

Example OUTPUT @Agb2200;":ROUT:COUP:STAT ALL,ON" OUTPUT @Agb2200;":ROUT:COUP:STAT? 2" ENTER @Agb2200;A

In this example, A will be 1.

OUTPUT @Agb2200;":ROUT:COUP:PORT ALL,'1'" OUTPUT @Agb2200;":ROUT:COUP:STAT ALL,ON" OUTPUT @Agb2200;":ROUT:CLOS (@10103)"

This example controls the card 1 relays and connects the inputs 1 and 2 to the outputs 3 and 4 respectively. :ROUT:CLOS (@10104) will make the same result.

SCPI Command Reference · Parameter Explanation

Page 166 of the Agilent Technologies B2201A user manual — [:ROUTe]:OPEN:CARD
Page 166 · click to enlarge
Page 166

SCPI Command Reference [:ROUTe]:FUNCtion

[:ROUTe]:FUNCtion

This command sets the channel configuration, which determines how you specify the channel_list or card_number in other commands. The query returns the present channel configuration. At *RST, this parameter is set to ACONfig. See “Channel Configuration Mode” on page 3-13.

When configuration is changed by this command, all channels are opened (disconnected) and become same status as after :SYSTem:CPON ALL execution.

Syntax [:ROUTe]:FUNCtion channel_config

[:ROUTe]:FUNCtion?

channel_config ACONfigAuto Config Mode NCONfig : Normal Config Mode

Query response ACON or NCON

Example OUTPUT @Agb2200;":ROUT:FUNC ACON" OUTPUT @Agb2200;":ROUT:FUNC?" ENTER @Agb2200;A$

In this example, A$ will be ACON.

[:ROUTe]:OPEN:CARD

This command disconnects all input ports from all output ports for the specified card. For some modes (such as Single Route Mode, Couple Mode, Ground Mode, and Bias Mode), more complex disconnections may occur.

Syntax [:ROUTe]:OPEN:CARD card_number

card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL

Example OUTPUT @Agb2200;":ROUT:OPEN:CARD ALL"

Parameter Explanation · Parameter Explanation

Page 167 of the Agilent Technologies B2201A user manual — [:ROUTe]:OPEN[:LIST]
Page 167 · click to enlarge
Page 167

[:ROUTe]:OPEN[:LIST]

[:ROUTe]:OPEN[:LIST]

This command disconnects the input ports from output ports as specified in channel_list. The query returns 0 or 1 for each channel specified by channel_list, in same order as specified by channel_list.

For some modes (such as Single Route Mode, Couple Mode, Ground Mode, and Bias Mode), more complex disconnections may occur.

Syntax [:ROUTe]:OPEN[:LIST] (@channel_list)

[:ROUTe]:OPEN[:LIST]? (@channel_list)

channel_list Channels to open (disconnect) or to check. For channel_list, see “Switch Control” on page 4-5.

Query response open_status{, open_status}

1. opened 0 : closed

Example OUTPUT @Agb2200;":ROUT:OPEN:CARD ALL" OUTPUT @Agb2200;":ROUT:CLOS (@10101,10202)" OUTPUT @Agb2200;":ROUT:OPEN? (@10101,10102,10201,10202)" ENTER @Agb2200;A$

In this example, A$ will be 0,1,1,0.

SCPI Command Reference · Parameter Explanation

Page 168 of the Agilent Technologies B2201A user manual — [:ROUTe]:SYMBol:CHANnel
Page 168 · click to enlarge
Page 168

SCPI Command Reference [:ROUTe]:SYMBol:CHANnel

[:ROUTe]:SYMBol:CHANnel

In the GPIB local mode, the B2200 uses 01 to 48 (two digits numbers) to specify the output ports (channels) 1 to 48 respectively. This command specifies a symbol string and lets the B2200 use the specified string instead of the two digits number. The query returns the symbol string.

Maximum value of the two digits number depends on the B2200 configuration. It is 12 for the normal configuration mode or the auto configuration mode with one matrix card, 24 for the auto configuration mode with two matrix cards, 36 for the auto configuration mode with three matrix cards, or 48 for the auto configuration mode with four matrix cards.

Reboot, *RST, and :SYSTem:CPON clear the symbol strings. After that, the initial symbol string (two digits numbers 01 to 48) will be set.

Syntax [:ROUTe]:SYMBol:CHANnel card_number,channel_number,'string'

[:ROUTe]:SYMBol:CHANnel? card_number,channel_number

card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL ALL is not available for query.

channel_number Output port number: 1 to 48 for Auto, 1 to 12 for Normal.

string Symbol string set to the specified output port (channel). Maximum 6 characters.

Numeric characters, alphabetic characters; upper case and lower case, space, and the following symbols are available.

! " # $ % & ’ ( ) * + , - . / > ? : ~ ; < = > ? @ [ \ ] ‘ { | } ^ _

Query response string

Example OUTPUT @Agb2200;":ROUT:SYMB:CHAN 0,10,'GATE'" OUTPUT @Agb2200;":ROUT:SYMB:CHAN? 0,10" ENTER @Agb2200;A$

This example assigns the string GATE to the output port 10. After that, GATE is used instead of 10. In this example, A$ will be GATE.

Parameter Explanation

Page 169 of the Agilent Technologies B2201A user manual — [:ROUTe]:SYMBol:PORT
Page 169 · click to enlarge
Page 169

[:ROUTe]:SYMBol:PORT

[:ROUTe]:SYMBol:PORT

In the GPIB local mode, the B2200 uses 01 to 14 (two digits numbers) to specify the input ports 1 to 14 respectively. This command specifies a symbol string and lets the B2200 use the specified string instead of the two digits number. The query returns the symbol string.

Reboot and *RST clear the symbol strings. After that, the initial symbol string (two digits numbers 01 to 14) will be set.

:SYSTem:CPON does not clear the symbol strings.

Syntax [:ROUTe]:SYMBol:PORT port_number,'symbol_string'

[:ROUTe]:SYMBol:PORT? port_number

port_number Input port number: 1 to 14.

symbol_string Symbol string set to the specified input port. Maximum 6 characters.

Numeric characters, alphabetic characters; upper case and lower case, space, and the following symbols are available.

! " # $ % & ’ ( ) * + , - . / > ? : ~ ; < = > ? @ [ \ ] ‘ { | } ^ _

Query response symbol_string

Example OUTPUT @Agb2200;":ROUT:SYMB:PORT 1,'SMU1'" OUTPUT @Agb2200;":ROUT:SYMB:PORT? 1" ENTER @Agb2200;A$

This example assigns the string SMU1 to the input port 1. After that, SMU1 is used instead of 01. In this example, A$ will be SMU1.

SCPI Command Reference · Parameter Explanation

Page 170 of the Agilent Technologies B2201A user manual — :SYSTem:CCONfig?
Page 170 · click to enlarge
Page 170

SCPI Command Reference :SYSTem:BEEP

:SYSTem:BEEP

This command enables/disables the beeper.

Syntax :SYSTem:BEEP state

state ON or 1enables the beeper OFF or 0 : disables the beeper

Example OUTPUT @Agb2200;":SYST:BEEP ON"

:SYSTem:CCONfig?

This query command is just to keep compatibility with the Agilent E5250A Low Leakage Switch Mainframe. This query command returns the card configuration information.

Syntax :SYSTem:CCONfig? card_number

card_number card number1 , 2 , 3 , or 4

Query response card_configuration

Always returns #10.

Example OUTPUT @Agb2200;":SYST:CCON? 1" ENTER @Agb2200;A$ PRINT "Card configuration = ";A$

Parameter Explanation · Parameter Explanation

Page 171 of the Agilent Technologies B2201A user manual — :SYSTem:CDEScription?
Page 171 · click to enlarge
Page 171

:SYSTem:CDEScription?

:SYSTem:CDEScription?

This query command returns a description of the specified card.

Syntax :SYSTem:CDEScription? card_number

card_number For Auto Config mode0 For Normal Config mode: 1 , 2 , 3 , or 4

Query response "card_description"

If the designated card is not installed or initialize failed, this command returns "No Card".

For B2200A:

  • In the Auto Config mode, " card_description " response is one of following:
  • "B2210A 14x12 Femto Leakage Switch Module"
  • "B2210A 14x24 Femto Leakage Switch Module"
  • "B2210A 14x36 Femto Leakage Switch Module"
  • "B2210A 14x48 Femto Leakage Switch Module"
  • In the Normal Config mode:

"B2210A 14x12 Femto Leakage Switch Module"

For B2201A:

  • In the Auto Config mode, " card_description " response is one of following:
  • "B2211A 14x12 Low Leakage Switch Module"
  • "B2211A 14x24 Low Leakage Switch Module"
  • "B2211A 14x36 Low Leakage Switch Module"
  • "B2211A 14x48 Low Leakage Switch Module"
  • In the Normal Config mode:

"B2211A 14x12 Low Leakage Switch Module"

Example OUTPUT @Agb2200;":SYST:CDES? 1" ENTER @Agb2200;A$

SCPI Command Reference · Parameter Explanation

Page 172 of the Agilent Technologies B2201A user manual — :SYSTem:CPON
Page 172 · click to enlarge
Page 172

SCPI Command Reference :SYSTem:CPON

:SYSTem:CPON

This command resets the specified card. Status after this command is shown below:

Channel Configuration: Not changed.

Connection RuleFree
Connection SequenceBreak Before Make
Bias ModeOff
Bias Input Port10

Bias-enabled Channels: All channels (output ports) are bias-enabled.

Couple ModeOff

Couple Input Port: Cleared.

Ground ModeOff
Ground Input Port12

Ground-enabled Channels: Cleared.

Ground-enabled Ports: Cleared.

Card Channel Status: All relays are opened.

Self-Test Result: Not changed.

Input Port Symbol String: Not changed.

Output Port Symbol String: Cleared.

For the device reset state, see *RST.

Syntax :SYSTem:CPON card_number

card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL

Example OUTPUT @Agb2200;":SYST:CPON ALL"

Parameter Explanation

Page 173 of the Agilent Technologies B2201A user manual — :SYSTem:DISPlay:LCD
Page 173 · click to enlarge
Page 173

:SYSTem:CTYPe?

:SYSTem:CTYPe?

This query command returns the ID of the specified card.

Syntax :SYSTem:CTYPe? card_number

card_number For Auto Config mode0 For Normal Config mode: 1 , 2 , 3 , or 4

Query response AGILENT TECHNOLOGIES,model,0,revision

model = B2210A or B2211A

revision = 1 or later version

If designated card is not installed or initialize failed, this command returns NONE,NONE,0,0.

Example OUTPUT @Agb2200;":SYST:CTYPE? 1" ENTER @Agb2200;A$

:SYSTem:DISPlay:LCD

This command enables/disables the front panel Liquid Crystal Display when the B2200 is in the GPIB remote mode. The front panel LCD will be automatically set to ON by the mode transition from GPIB remote to GPIB local.

Syntax :SYSTem:DISPlay:LCD state

state ON or 1enables the LCD OFF or 0 : disables the LCD (initial setting)

Even though state is OFF, some status indicators on the LCD will be effective.

Example OUTPUT @Agb2200;":SYST:DISP:LCD 1"

SCPI Command Reference · Parameter Explanation · Parameter Explanation

Page 174 of the Agilent Technologies B2201A user manual — :SYSTem:DISPlay:STRing
Page 174 · click to enlarge
Page 174

SCPI Command Reference :SYSTem:DISPlay:LED

:SYSTem:DISPlay:LED

This command enables/disables the front panel Light Emitting Diodes matrix when the B2200 is in the GPIB remote mode. The front panel LED matrix will be automatically set to ON by the mode transition from GPIB remote to GPIB local.

Syntax :SYSTem:DISPlay:LED state

state ON or 1enables the LEDs (initial setting) OFF or 0 : disables the LEDs

Example OUTPUT @Agb2200;":SYST:DISP:LED 1"

:SYSTem:DISPlay:STRing

When the B2200 is in the GPIB remote mode, you can display a string (maximum 39 characters) on the front panel LCD. This command specifies the string and displays it on the LCD.

The specified string data is only displayed on the front panel LCD. It is not memorized. So a display-change operation will clear and delete the string data.

Syntax :SYSTem:DISPlay:STRing 'string'

string String to be displayed on the front panel LCD. Maximum 38 characters.

Numeric characters, alphabetic characters; upper case and lower case, space, and the following symbols are available.

! " # $ % & ’ ( ) * + , - . / > ? : ~ ; < = > ? @ [ \ ] ‘ { | } ^ _

Example OUTPUT @Agb2200;":SYST:DISP:STR 'E:SMU1,B:SMU2,C:SMU3'"

This example displays E:SMU1,B:SMU2,C:SMU3 on the front panel LCD.

Parameter Explanation · Parameter Explanation

Page 175 of the Agilent Technologies B2201A user manual — :SYSTem:KLC
Page 175 · click to enlarge
Page 175

:SYSTem:ERRor?

:SYSTem:ERRor?

This query command reads the error from the head of the error queue and removes that error from the queue.

Syntax :SYSTem:ERRor?

Query response err_no,"message"

err_no is numeric response data, and message is string response data.

If there has been no error (error queue is empty), the response to this query is as follows:

Example OUTPUT @Agb2200;":SYST:ERR?" ENTER @Agb2200;A,B$

:SYSTem:KLC

This command locks/unlocks the front panel keys.

Syntax :SYSTem:KLC state

state ON or 1lock
OFF or 0unlock

Example OUTPUT @Agb2200;":SYST:KLC ON"

SCPI Command Reference · 0,"No error" · Parameter Explanation

Page 176 of the Agilent Technologies B2201A user manual — :SYSTem:MEMOry:DELete
Page 176 · click to enlarge
Page 176

SCPI Command Reference :SYSTem:MEMOry:COMMent

:SYSTem:MEMOry:COMMent

This command memorizes the comment for the B2200 setup information specified by memory_number. The previous comment will be deleted. The query returns the comment for the specified setup.

Syntax :SYSTem:MEMOry:COMMent memory_number,'comment'

:SYSTem:MEMOry:COMMent? memory_number

memory_number memory number: 1 to 8.

comment Comment. Maximum 16 characters. If this command is entered with empty comment, the comment will be cleared.

Numeric characters, alphabetic characters; upper case and lower case, space, and the following symbols are available.

! " # $ % & ’ ( ) * + , - . / > ? : ~ ; < = > ? @ [ \ ] ‘ { | } ^ _

Example OUTPUT @Agb2200;":SYST:MEMO:COMM 1,'1-1,2-13,3-25'" OUTPUT @Agb2200;":SYST:MEMO:COMM? 1" ENTER @Agb2200;A$

In this example, A$ will be 1-1,2-13,3-25.

:SYSTem:MEMOry:DELete

This command deletes the B2200 setup information and the comment specified by memory_number.

Syntax :SYSTem:MEMOry:DELete memory_number

memory_number memory number: 1 to 8.

Example OUTPUT @Agb2200;":SYST:MEMO:DEL 1"

Parameter Explanation · Parameter Explanation

Page 177 of the Agilent Technologies B2201A user manual — :SYSTem:MEMOry:SAVE
Page 177 · click to enlarge
Page 177

:SYSTem:MEMOry:LOAD

:SYSTem:MEMOry:LOAD

This command restores the B2200 setup information specified by memory_number. It must be already saved by “:SYSTem:MEMOry:SAVE”.

Syntax :SYSTem:MEMOry:LOAD memory_number

memory_number memory number: 1 to 8.

Example OUTPUT @Agb2200;":SYST:MEMO:SAVE 1" OUTPUT @Agb2200;":SYST:MEMO:LOAD 1"

:SYSTem:MEMOry:SAVE

This command saves the present setup information of the B2200 into the internal memory specified by memory_number. The previous setup will be deleted. To memorize the comment for the setup information, use “:SYSTem:MEMOry:COMMent”.

The following setup information will be saved.

  • Configuration mode
  • Connection rule, connection sequence
  • Bias mode status, bias ports, bias-enabled channels
  • Couple mode status, couple ports
  • Ground mode status, ground ports, ground-enabled channels, ground-enabled ports
  • Connection status of all switches
  • Symbol string of all input ports and output ports

Syntax :SYSTem:MEMOry:SAVE memory_number

memory_number memory number: 1 to 8.

Example OUTPUT @Agb2200;":SYST:MEMO:SAVE 1"

SCPI Command Reference · Parameter Explanation · Parameter Explanation

Page 178 of the Agilent Technologies B2201A user manual — :SYSTem:VERSion?
Page 178 · click to enlarge
Page 178

SCPI Command Reference :SYSTem:PEN

:SYSTem:PEN

This command enables/disables the light pen.

Syntax :SYSTem:PEN state

state ON or 1enables the light pen OFF or 0 : disables the light pen

Example OUTPUT @Agb2200;":SYST:PEN ON"

:SYSTem:VERSion?

This query command returns the SCPI version number for which the B2200 complies.

Syntax :SYSTem:VERSion?

Query response YYYY.V

YYYY is the year (for example, 2004).

V is the approved revision number for that year.

Example OUTPUT @Agb2200;":SYST:VERS?" ENTER @Agb2200;A$

Parameter Explanation

Page 179 of the Agilent Technologies B2201A user manual — Status Reporting Structure
Page 179 · click to enlarge
Page 179

Status Reporting Structure

This section describes the status reporting structure used in the B2200. These are IEEE 488.2 status structures that can be set and read by the SCPI Common Commands as described in “Common Commands” on page 5-4.

Status Reporting Structure

The status reporting structure consists of the following:

  • Status Byte Register
  • Service Request “Enable” Register
  • Standard Event Status Register
  • Standard Event Status “Enable” Register
  • Output Queue

In general, the status reporting structure is used to request service via Service Request (SRQ) or to indicate a specific condition, such as operation complete.

See the following figures for a model of the B2200 status reporting structure.

SCPI Command Reference · Status Reporting Structure

Page 180 of the Agilent Technologies B2201A user manual — Status Reporting Structure
Page 180 · click to enlarge
Page 180

SCPI Command Reference Status Reporting Structure

Figure 5-1 Status Reporting Structure of B2200

Agilent Technologies B2201A — Status Reporting Structure diagram (page 180)
Fig. 20 · Status Reporting Structure

The above figure shows Service Request Generation. When a condition occurs that requires service, the instrument sets Request Service bit (RQS − bit6) of Status Byte, and sends a Service Request (SRQ) via GPIB bus to the controller. So, the controller can execute an interrupt service routine (must be in the program) that uses a Serial Poll to read Status Byte of each instrument to determine which instrument requested service. See “Status Byte Register” on page 5-55.

Page 181 of the Agilent Technologies B2201A user manual — Status Byte Register
Page 181 · click to enlarge
Page 181

Status Byte Register

The B2200 Status Byte Register contains bits (ESB and MAV) for status summary messages from other registers. The status of these bits depends on the condition of the Standard Event Status Register and the Output Queue. If service request occurs, Bit6 (RQS) of Status Byte Register is set.

See Figure 5-2 and Table 5-1 for the B2200 Status Byte Register.

Figure 5-2 Status Byte Register of B2200

Agilent Technologies B2201A — Status Byte Register diagram (page 181)
Fig. 21 · Status Byte Register

The Status Byte Register can be read with either a serial poll or the *STB? common query command (see “Common Commands” on page 5-4).

Serial poll is a low-level GPIB command that can be executed by SPOLL in HP BASIC, such as the following:

Both serial poll and STB? read the Status Byte Register identically, but have following difference:

  • SPOLL returns RQS for bit6
  • *STB? returns MSS for bit6

RQS and MSS are always the same value, so returned Status Byte value is always the same for these two methods.

SCPI Command Reference · Status Byte Register · Status=SPOLL(@Agb2200)

Page 182 of the Agilent Technologies B2201A user manual — Status Byte Register
Page 182 · click to enlarge
Page 182

SCPI Command Reference Status Byte Register

In general, use serial polling (not *STB?) inside interrupt service routines. Use *STB? in other cases (not in interrupt service routine) when you want to know the value of Status Byte.

Table 5-1 Status Byte Register of B2200

0 Not Used. Always zero.

1 Not Used. Always zero.

2 Not Used. Always zero.

3 Not Used. Always zero.

4 Message Available (MAV) Queue Summary Message

Indicates whether Output Queue is empty. If unread message exists, this bit is set to 1. Otherwise, this bit is set to 0. See “Output Queue” on page 5-60.

5 Standard Event Status Bit (ESB) Summary Message

Indicates whether one or more of the enabled Standard Events Status Register bits is set.

6 Request Service (RQS) Message

Indicates whether a SRQ (Service Request) has occurred. Read by Serial Poll.

Master Summary Status (MSS) Summary Message

Indicates that the instrument has at least one reason for requesting service. Read by *STB?.

7 Not Used. Always zero.

Bit Definition Explanation

Page 183 of the Agilent Technologies B2201A user manual — Service Request Enable Register
Page 183 · click to enlarge
Page 183

SCPI Command Reference Service Request Enable Register

Service Request Enable Register

The Service Request Enable Register is an 8-bit register that can be used by the programmer to select which summary messages in the Status Byte Register may cause service requests. See Figure 5-3.

Figure 5-3 Service Request Enable Register of B2200

Agilent Technologies B2201A — Service Request Enable Register diagram (page 183)
Fig. 22 · Service Request Enable Register
Page 184 of the Agilent Technologies B2201A user manual — Standard Event Status Register
Page 184 · click to enlarge
Page 184

SCPI Command Reference Standard Event Status Register

Standard Event Status Register

The Standard Event Status Register has specific events assigned to specific bits. See Figure 5-4 and Table 5-2.

Figure 5-4 Standard Event Status Register of B2200

Agilent Technologies B2201A — Standard Event Status Register diagram (page 184)
Fig. 23 · Standard Event Status Register
Page 185 of the Agilent Technologies B2201A user manual — Standard Event Status Enable Register
Page 185 · click to enlarge
Page 185

SCPI Command Reference Standard Event Status Enable Register

Table 5-2 Standard Event Status Register of B2200

0 Operation Complete (OPC)

This event bit has meaning only if a request to monitor is set by the *OPC command. See “*OPC” on page 5-8. This bit is set to 1 if there are no pending operations.

1 Not Used Always 0.

2 Query Error (QYE) • An attempt is being made to read data from the Output Queue when no data is present or pending.

  • Data in the Output Queue has been lost.

3 Device Dependent Error (DDE)

This event bit indicates that an error has occurred which is not a Command Error, a Query Error, or an Execution Error.

4 Execution Error (EXE) Syntax of command is correct, but cannot be executed due to some condition of the B2200.

5 Command Error (CME)

A command syntax error has been detected.

6 Not Used Always 0.

7 Power On (PON) This event bit indicates that an off-to-on transition has occurred in instrument's power supply.

8 to 15 Not used Always 0.

Standard Event Status Enable Register

The Standard Event Status “Enable” Register is an 8-bit register that can be used by the programmer to select which bits of Standard Event Status Register are enabled. The status of the enabled bits are ORed together, and result of OR will be reported to the ESB bit (Bit5) of the Status Byte Register.

The 8 bits of this register correspond to the 8 bits of the Standard Event Status Register. See Figure 5-4.

Bit Definition Explanation

Page 186 of the Agilent Technologies B2201A user manual — Output Queue
Page 186 · click to enlarge
Page 186

SCPI Command Reference Output Queue

Output Queue

The Output Queue stores response messages until they are read. If an unread message exists, Bit4 (Message Available − MAV) of the Status Byte Register is set to 1. So, Bit4 is used to synchronize information exchange with the controller. See Figure 5-5.

Figure 5-5 Output Queue of B2200

Agilent Technologies B2201A — Output Queue diagram (page 186)
Fig. 24 · Output Queue

The Response Formatter places Data Byte Messages and END messages into the Output Queue in response to query commands. These messages are removed from the Output Queue as they are read by the controller. As long as the Output Queue contains an unread message, MAV is 1.

Page 187 of the Agilent Technologies B2201A user manual — 6 VXI plug&play Driver
Page 187 · click to enlarge
Page 187

6 VXI plug&play Driver

Page 188 of the Agilent Technologies B2201A user manual — 6 VXI plug&play Driver
Page 188 · click to enlarge
Page 188

This chapter introduces the VXI plug&play driver available for Agilent B2200, and consists of the following sections:

  • “System Requirements”
  • “Installing VXIplug&play Driver”
  • “Driver Functions”

VXI plug&play Driver

Page 189 of the Agilent Technologies B2201A user manual — System Requirements
Page 189 · click to enlarge
Page 189

VXI plug&play Driver System Requirements

System Requirements

The following system environments are required.

  • Operating System

Microsoft Windows XP Professional, Windows 2000, Windows NT 4.0, or Windows 95. It must be supported by the application development environment.

  • Application Development Environment (programming environment)

Microsoft Visual Basic, Microsoft Visual C++, Borland C++Builder, National Instruments LabWindows or LabVIEW, or Agilent VEE.

  • Agilent T&M Programmers Toolkit for Visual Studio .NET

Agilent W1140A or equivalent. Needed for Visual Basic .NET users.

  • GPIB (IEEE 488) Interface and 32-bit VISA I/O Library

Agilent 82357A USB/GPIB interface, E5810A LAN/GPIB gateway, 82350B GPIB interface, or equivalent. These models include Agilent VISA and SICL I/O libraries.

  • Computer and peripherals

Required specifications depend on the application development environment. See manual of the application development environment. The flexible disk drive is required to install the VXIplug&play driver. The CD-ROM drive is required to install the software needed to use driver.

  • Minimum disk space

1 MB for the Agilent B2200 VXIplug&play driver

Page 190 of the Agilent Technologies B2201A user manual — Installing VXI plug&play Driver
Page 190 · click to enlarge
Page 190

VXI plug&play Driver Installing VXIplug&play Driver

Installing VXI plug&play Driver

The installation flow is shown below. If you have already installed the GPIB interface card, VISA I/O library, and programming software on your PC, skip steps 1 through 5.

1. Install the GPIB interface card into your PC.

See the interface card manual. Note the model number of the interface card, as you may need it to configure the interface (in step 3).

2. Install VISA I/O library.

Follow the instructions in the I/O library’s setup program.

3. Configure and check the GPIB interface.

See the I/O library manual.

4. Install the programming software.

Follow the setup program instructions.

5. If you use Microsoft Visual Basic .NET, install the Agilent T&M Programmers Toolkit.
6. Install the VXI plug&play driver as shown below.

a. Insert the Agilent B2200 Plug&Play Driver Disk into the flexible disk drive connected to your PC.

b. Execute the SETUP.EXE program on the diskette. The program

automatically installs the following files.

Directory\Program Files\VISA\win xx \agb220xa ( xx depends on OS)
  • agb220xa.bas
  • agb220xa.c
  • agb220xa.def
  • agb220xa.fp
  • agb220xa.GID
  • agb220xa.h
  • agb220xa.hlp
  • readme.txt
  • DelsL1.isu
Page 191 of the Agilent Technologies B2201A user manual — Driver Functions
Page 191 · click to enlarge
Page 191

Driver Functions

Table 6-1 lists all the functions for the Agilent B2200. You will see a brief description of the functions in the table.

For the description, syntax and parameters of the function, refer to the reference section following this table. The driver functions in the reference section will appear in the alphabetical order.

Table 6-1 Agilent B2200 Driver Function Lists

Miscellaneous agb220xa_init Initializes the B2200. agb220xa_close Closes the connection with the B2200. agb220xa_reset Executes the B2200 reset. agb220xa_self_test Executes the B2200 self-test. agb220xa_error_query Queries for the B2200 error code/message. agb220xa_error_message Queries for the driver error. agb220xa_revision_query Queries for the B2200 firmware/driver revisions. agb220xa_timeOut Sets the timeout. agb220xa_timeOut_Q Queries for the timeout setting. agb220xa_errorQueryDetect Sets the automatic error checking. agb220xa_errorQueryDetect_Q Queries for the automatic error checking setting. agb220xa_dcl Sends the Device Clear. agb220xa_esr_Q Queries for the ESR status. agb220xa_readStatusByte_Q Reads the B2200 status byte. agb220xa_opc_Q Checks the B2200 operation completion status. Mode Control agb220xa_func Sets the configuration mode. agb220xa_connRuleSeq Sets the connection rule/sequence. Couple Mode agb220xa_couplePort Selects the couple ports of the specified card. agb220xa_coupleState Sets the couple mode ON or OFF for the specified card. agb220xa_detectCouplePort Detects and sets the couple ports for all cards. Bias Mode agb220xa_biasPort Selects the input bias port of the specified card. agb220xa_biasChanCard Bias-enables all output ports of the specified card. agb220xa_biasChanList Bias-enables all output ports of the specified channel list. agb220xa_biasState Sets the bias mode ON or OFF for the specified card. agb220xa_biasChanList_Q Queries for the bias channel list.

VXI plug&play Driver · Driver Functions · Category Function Description

Page 192 of the Agilent Technologies B2201A user manual — Driver Functions
Page 192 · click to enlarge
Page 192

VXI plug&play Driver Driver Functions

Ground Mode agb220xa_groundPort Selects the input ground port of the specified card. agb220xa_groundChanCard Ground-enables all output ports of the specified card. agb220xa_groundChanList Ground-enables all output ports of the specified channel list. agb220xa_unusedPort Ground-enables the specified input ports of the specified card. agb220xa_groundState Sets the ground mode ON or OFF for the specified card. agb220xa_groundChanList_Q Queries for the ground channel list. Route Control agb220xa_closeList Closes the channel list. agb220xa_openList Opens the channel list. agb220xa_openCard Opens all output on the card. agb220xa_closeList_Q Queries for the channel list status. agb220xa_openList_Q agb220xa_closeCard_Q Queries for the closed channel list on the card. C/G Compensation

agb220xa_compenC Performs the compensation for the C/G data measured by the Agilent 4284A. agb220xa_selectCompenFile Selects the compensation data file used by the agb220xa_compenC function. Diagnostics agb220xa_testExec_Q Performs diagnostics and returns the test result. agb220xa_testClear Clears the specified test result. Passthrough Functions

agb220xa_cmd Sends a command. agb220xa_cmdInt Sends a command with an integer parameter. agb220xa_cmdReal Sends a command with a real parameter. agb220xa_cmdData_Q Sends a command to read any data. agb220xa_cmdString_Q Sends a command to read string response. agb220xa_cmdInt16_Q Sends a command to read 16 bit integer response. agb220xa_cmdInt16Arr_Q Sends a command to read 16 bit integer array response. agb220xa_cmdInt32_Q Sends a command to read 32 bit integer response. agb220xa_cmdInt32Arr_Q Sends a command to read 32 bit integer array response. agb220xa_cmdReal64_Q Sends a command to read 64 bit real response. agb220xa_cmdReal64Arr_Q Sends a command to read 64 bit real array response.

Category Function Description

Page 193 of the Agilent Technologies B2201A user manual — agb220xa_biasChanList
Page 193 · click to enlarge
Page 193

agb220xa_biasChanCard

This function bias-enables or disables all the output ports of the specified card.

Syntax ViStatus _VI_FUNC agb220xa_biasChanCard(ViSession vi, ViInt16 disable_enable, ViInt16 bias_cardno);

Parameters vi Instrument handle returned from agb220xa_init( ).

disable_enable Status of the card, bias enabled or bias disabled.

0. sets bias enabled card.
1. sets bias disabled card.

bias_cardno Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5 (all card) in the normal configuration mode, or 0 (all card in the auto configuration mode). For the configuration mode, see agb220xa_func.

agb220xa_biasChanList

This function bias-enables or disables all the output ports specified by the biaschan_list.

The parameter “biaschan_list” is an array of integers with each integer representing one channel. The last number of the “biaschan_list” should be “0” (numeric zero) to identify the end of the list. The maximum number of channels that can be specified by the list is 100.

Syntax ViStatus _VI_FUNC agb220xa_biasChanList(ViSession vi, ViInt16 biaschan_disen, ViInt32_VI_FAR biaschan_list[ ] );

Parameters vi Instrument handle returned from agb220xa_init( ).

biaschan_disen Status of the port, bias enabled or bias disabled.

0. sets bias enabled port.
1. sets bias disabled port.

biaschan_list[ ] Channel numbers. 5 digits integer. ABCDE. where A: card number, BC: input port number, DE: output port number. Top zero(s) can be ignored. For example, if A=0, BC=01, and DE=01, channel number should be 101 instead of 00101.

VXI plug&play Driver · Driver Functions

Page 194 of the Agilent Technologies B2201A user manual — agb220xa_biasChanList_Q
Page 194 · click to enlarge
Page 194

VXI plug&play Driver Driver Functions

agb220xa_biasChanList_Q

This function will query the instrument for the status, bias enabled or bias disabled, for the channels given in the list.

The parameter “biaschan_list” is an array of integers with each integer representing one channel. The last number of the “biaschan_list” should be “0” (numeric zero) to identify the end of the list. The maximum number of channels that can be specified by the list is 100.

The “bias_status” parameter is an array of integers containing the return values of the query. The “bias_status” array returned will correspond one to one with “biaschan_list” parameter.

Syntax ViStatus _VI_FUNC agb220xa_biasChanList_Q(ViSession vi, ViInt16 bias_disen, ViInt32_VI_FAR biaschan_list[ ], ViInt32_VI_FAR bias_status[ ] );

Parameters vi Instrument handle returned from agb220xa_init( ).

bias_disen Query type.

0. checks if the port is the bias enabled.
1. checks if the port is the bias disabled.

biaschan_list[ ] Channel numbers to check the status. 5 digits integer. ABCDE. where A: card number, BC: input port number, DE: output port number. Top zero(s) can be ignored. For example, if A=0, BC=01, and DE=01, channel number should be 101 instead of 00101.

bias_status[ ] Status of the channels given in the biaschan_list. Returned value depends on the setting of bias_disen as shown below:

when bias_disen=0, 0 means bias disabled, 1 means enabled.

when bias_disen=1, 0 means bias enabled, 1 means disabled.

Page 195 of the Agilent Technologies B2201A user manual — agb220xa_biasState
Page 195 · click to enlarge
Page 195

agb220xa_biasPort

This function will select which input port is the bias port on the specified card. For each card, you can specify the same or different bias port.

Syntax ViStatus _VI_FUNC agb220xa_biasPort(ViSession vi, ViInt16 bport_cardno, ViInt16 bias_port);

Parameters vi Instrument handle returned from agb220xa_init( ).

bport_cardno Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5 (all card) in the normal configuration mode, or 0 (all card in the auto configuration mode). For the configuration mode, see agb220xa_func.

bias_port Input port number to be set to the bias port. 1 to 14 (input port 1 to input port 14).

If the bias input port and a couple port have been assigned to the same input port, the bias mode and the couple mode cannot be used in parallel.

agb220xa_biasState

This function controls the bias mode for the specified card. When bias mode is ON, the input bias port is connected to all bias enabled output ports that are not connected to any other input ports. Bias disabled output ports are never connected to the input bias port when bias mode is ON.

The bias mode cannot be set to ON when the ground mode is ON.

Syntax ViStatus _VI_FUNC agb220xa_biasState(ViSession vi, ViInt16 bstate_cardno, ViInt16 state);

Parameters vi Instrument handle returned from agb220xa_init( ).

bstate_cardno Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5 (all card) in the normal configuration mode, or 0 (all card in the auto configuration mode). For the configuration mode, see agb220xa_func.

state Bias mode. 0 (OFF) or 1 (ON).

VXI plug&play Driver · Driver Functions

Page 196 of the Agilent Technologies B2201A user manual — agb220xa_closeCard_Q
Page 196 · click to enlarge
Page 196

VXI plug&play Driver Driver Functions

agb220xa_close

This function terminates the software connection to the instrument and deallocates system resources. It is generally a good programming habit to close the instrument handle when the program is done using the instrument.

Syntax ViStatus _VI_FUNC agb220xa_close(ViSession vi);

Parameters vi Instrument handle returned from agb220xa_init( ).

agb220xa_closeCard_Q

This function will query the card for the channels closed of the specified card.

The parameter “closechan_list” contains the channel numbers returned by the instrument. This will be an array of integers terminated by “zero” to identify the end of the list. Array of enough length should be passed to the function.

Syntax ViStatus _VI_FUNC agb220xa_closeCard_Q(ViSession vi, ViInt16 close_card, ViInt32_VI_FAR closechan_list[ ] );

Parameters vi Instrument handle returned from agb220xa_init( ).

close_card Card number. 1 (card 1), 2 (card 2), 3 (card 3), or 4 (card 4) in the normal configuration mode, or 0 (all card in the auto configuration mode). For the configuration mode, see agb220xa_func.

closechan_list[ ] Channels closed of the specified card.

Page 197 of the Agilent Technologies B2201A user manual — agb220xa_closeList
Page 197 · click to enlarge
Page 197

agb220xa_closeList

This function will connect the input ports to the output ports specified by the channel list.

The parameter “closechan_list” is an array of integers with each integer representing one channel. The last number of the “closechan_list” should be “0” (numeric zero) to identify the end of the list. The maximum number of channels that can be specified by the list is 100.

Syntax ViStatus _VI_FUNC agb220xa_closeList(ViSession vi, ViInt32_VI_FAR closechan_list[ ] );

Parameters vi Instrument handle returned from agb220xa_init( ).

closechan_list[ ] Channel numbers to connect. 5 digits integer. ABCDE. where A: card number, BC: input port number, DE: output port number. Top zero(s) can be ignored. For example, if A=0, BC=01, and DE=01, channel number should be 101 instead of 00101.

VXI plug&play Driver · Driver Functions

Page 198 of the Agilent Technologies B2201A user manual — agb220xa_cmd
Page 198 · click to enlarge
Page 198

VXI plug&play Driver Driver Functions

agb220xa_closeList_Q

This function will query the instrument for the channels closed given in the “closechan_list”.

The parameter “closechan_list” is an array of integers with each integer representing one channel. The last number of the “closechan_list” should be “0” (numeric zero) to identify the end of the list. The maximum number of channels that can be specified by the list is 100.

The “close_status” parameter is an array of integers containing the return values of the query. The “close_status” array returned will correspond one to one with “closechan_list” parameter.

Syntax ViStatus _VI_FUNC agb220xa_closeList_Q(ViSession vi, ViInt32_VI_FAR closechan_list[ ], ViInt32_VI_FAR close_status[ ] );

Parameters vi Instrument handle returned from agb220xa_init( ).

closechan_list[ ] Channel numbers to know the close status. 5 digits integer. ABCDE. where A: card number, BC: input port number, DE: output port number. Top zero(s) can be ignored. For example, if A=0, BC=01, and DE=01, channel number should be 101 instead of 00101.

close_status[ ] Status of the channels given in the closechan_list. 0 (opened) or 1 (closed).

agb220xa_cmd

This function passes the cmd_str string to the instrument. Must be a NULL terminated C string.

Syntax ViStatus _VI_FUNC agb220xa_cmd(ViSession vi, ViString cmd_str);

Parameters vi Instrument handle returned from agb220xa_init( ).

cmd_str Instrument command (cannot exceed 256 bytes in length).

Page 199 of the Agilent Technologies B2201A user manual — agb220xa_cmdInt
Page 199 · click to enlarge
Page 199

agb220xa_cmdData_Q

This function passes the cmd_str string to the instrument. This entry point will wait for a response which may be any data. You specify the cmd_str and size parameters, and get result[ ].

Syntax ViStatus _VI_FUNC agb220xa_cmdData_Q(ViSession vi, ViString cmd_str, ViInt32 size, ViChar_VI_FAR result[ ] );

Parameters vi Instrument handle returned from agb220xa_init( ).

cmd_str Instrument command (cannot exceed 256 bytes in length).

size Length of result in bytes. 2 to 32767.

result[ ] Response from instrument.

agb220xa_cmdInt

This function passes the cmd_str string to the instrument. This entry point passes the string in cmd_str followed by a space and then the integer in value. Note that either an Int16 or 32 can be passed as the Int16 will be promoted.

Syntax ViStatus _VI_FUNC agb220xa_cmdInt(ViSession vi, ViString cmd_str, ViInt32 value);

Parameters vi Instrument handle returned from agb220xa_init( ).

cmd_str Instrument command (cannot exceed 256 bytes in length).

value Parameter for command. -2147483647 to 2147483647.

VXI plug&play Driver · Driver Functions

Page 200 of the Agilent Technologies B2201A user manual — agb220xa_cmdInt16_Q
Page 200 · click to enlarge
Page 200

VXI plug&play Driver Driver Functions

agb220xa_cmdInt16Arr_Q

This function passes the cmd_str string to the instrument. This command expects a response that is a definite arbitrary block of 16 bit integers. You specify the cmd_str and size parameters, and get result[ ] and count.

Syntax ViStatus _VI_FUNC agb220xa_cmdInt16Arr_Q(ViSession vi, ViString cmd_str, ViInt32 size, ViInt16_VI_FAR result[ ], ViPInt32 count);

Parameters vi Instrument handle returned from agb220xa_init( ).

cmd_str Instrument command (cannot exceed 256 bytes in length).

size Size of result[ ] (number of items in the array). 1 to 2147483647.

result[ ] Response from instrument.

count Count of valid items in result[ ].

agb220xa_cmdInt16_Q

This function passes the cmd_str string to the instrument. This command expects a response that can be returned as a 16 bit integer.

Syntax ViStatus _VI_FUNC agb220xa_cmdInt16_Q(ViSession vi, ViString cmd_str, ViPInt16 result);

Parameters vi Instrument handle returned from agb220xa_init( ).

cmd_str Instrument command (cannot exceed 256 bytes in length).

result Response from instrument.

Page 201 of the Agilent Technologies B2201A user manual — agb220xa_cmdInt32_Q
Page 201 · click to enlarge
Page 201

agb220xa_cmdInt32Arr_Q

This function passes the cmd_str string to the instrument. This command expects a response that is a definite arbitrary block of 32 bit integers. You specify the cmd_str and size parameters, and get result[ ] and count.

Syntax ViStatus _VI_FUNC agb220xa_cmdInt32Arr_Q(ViSession vi, ViString cmd_str, ViInt32 size, ViInt32_VI_FAR result[ ], ViPInt32 count);

Parameters vi Instrument handle returned from agb220xa_init( ).

cmd_str Instrument command (cannot exceed 256 bytes in length).

size Size of result[ ] (number of items in the array). 1 to 2147483647.

result[ ] Response from instrument.

count Count of valid items in result[ ].

agb220xa_cmdInt32_Q

This function passes the cmd_str string to the instrument. This command expects a response that can be returned as a 32 bit integer.

Syntax ViStatus _VI_FUNC agb220xa_cmdInt32_Q(ViSession vi, ViString cmd_str, ViPInt32 result);

Parameters vi Instrument handle returned from agb220xa_init( ).

cmd_str Instrument command (cannot exceed 256 bytes in length).

result Response from instrument.

VXI plug&play Driver · Driver Functions

Page 202 of the Agilent Technologies B2201A user manual — agb220xa_cmdReal64Arr_Q
Page 202 · click to enlarge
Page 202

VXI plug&play Driver Driver Functions

agb220xa_cmdReal

This function passes the cmd_str string to the instrument. This entry point passes the string in cmd_str followed by a space and then the real in value. Note that either an Real32 or 64 can be passed as the Real32 will be promoted.

Syntax ViStatus _VI_FUNC agb220xa_cmdReal(ViSession vi, ViString cmd_str, ViReal64 value);

Parameters vi Instrument handle returned from agb220xa_init( ).

cmd_str Instrument command (cannot exceed 256 bytes in length).

value Parameter for command. -1E+300 to 1E+300.

agb220xa_cmdReal64Arr_Q

This function passes the cmd_str string to the instrument. This command expects a response that is a definite arbitrary block of 64 bit real. You specify the cmd_str and size parameters, and get result[ ] and count.

Syntax ViStatus _VI_FUNC agb220xa_cmdReal64Arr_Q(ViSession vi, ViString cmd_str, ViInt32 size, ViReal64_VI_FAR result[ ], ViPInt32 count);

Parameters vi Instrument handle returned from agb220xa_init( ).

cmd_str Instrument command (cannot exceed 256 bytes in length).

size Size of result[ ] (number of items in the array). 1 to 2147483647.

result[ ] Response from instrument.

count Count of valid items in result[ ].

Page 203 of the Agilent Technologies B2201A user manual — agb220xa_cmdString_Q
Page 203 · click to enlarge
Page 203

agb220xa_cmdReal64_Q

This function passes the cmd_str string to the instrument. This command expects a response that can be returned as a 64 bit real.

Syntax ViStatus _VI_FUNC agb220xa_cmdReal64_Q(ViSession vi, ViString cmd_str, ViPReal64 result);

Parameters vi Instrument handle returned from agb220xa_init( ).

cmd_str Instrument command (cannot exceed 256 bytes in length).

result Response from instrument.

agb220xa_cmdString_Q

This function passes the cmd_str string to the instrument. This entry point will wait for a response which must be a string (character data). You specify the cmd_str and size parameters, and get result[ ].

Syntax ViStatus _VI_FUNC agb220xa_cmdString_Q(ViSession vi, ViString cmd_str, ViInt32 size, ViChar_VI_FAR result[ ] );

Parameters vi Instrument handle returned from agb220xa_init( ).

cmd_str Instrument command (cannot exceed 256 bytes in length).

size Length of result in bytes. 2 to 32767.

result[ ] Response from instrument.

VXI plug&play Driver · Driver Functions

Page 204 of the Agilent Technologies B2201A user manual — agb220xa_connRuleSeq
Page 204 · click to enlarge
Page 204

VXI plug&play Driver Driver Functions

agb220xa_compenC

This function compensates capacitance/conductance data measured by the Agilent 4284A LCR meter, and returns compensation results. Before this function is executed, a compensation data file must be specified by using the agb220xa_selectCompenFile function. The file must contain the appropriate compensation coefficients for your measurement environment. For obtaining the compensation coefficients for your environment and creating the compensation data file, see “Capacitance Compensation” on page 4-20.

Syntax agb220xa_compenC(ViSession vi, ViReal64 frequency, ViReal64 raw_c, ViReal64 raw_g, ViPReal64 compen_c, ViPReal64 compen_g);

Parameters vi Instrument handle returned from agb220xa_init( ).

frequency Measurement frequency (in Hz). 1E3 (1 kHz) to 1E6 (1 MHz).

raw_c Capacitance (in F) measured by the 4284A.

raw_g Conductance (in S) measured by the 4284A.

compen_c Capacitance compensation result (in F). Returned value.

compen_g Conductance compensation result (in S). Returned value.

agb220xa_connRuleSeq

The function sets connection rule and connection sequence for the specified card.

Syntax ViStatus _VI_FUNC agb220xa_connRuleSeq(ViSession vi, ViInt16 cardno_ruleseq, ViInt16 rule, ViInt16 sequence);

Parameters vi Instrument handle returned from agb220xa_init( ).

cardno_ruleseq Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5 (all card) in the normal configuration mode, or 0 (all card in the auto configuration mode). For the configuration mode, see agb220xa_func.

rule Connection rule. 0 (free route) or 1 (single route).

sequence Connection sequence. 0, 1, or 2. See below.

1 (break before make)

2 (make before break)

0 (no sequence)

Page 205 of the Agilent Technologies B2201A user manual — agb220xa_couplePort
Page 205 · click to enlarge
Page 205

agb220xa_couplePort

This function sets the couple ports which are used for making kelvin connections on the specified card. The specified input port number will be coupled with the next input port and two output ports. For each card, you may setup the same or different couple ports. This command overwrites the previous couple port setting for the card.

The couple mode is controlled by the agb220xa_coupleState function.

If the bias input port and a couple port have been assigned to the same input port, the bias mode and the couple mode cannot be used in parallel.

If the ground input port and a couple port have been assigned to the same input port, the ground mode and the couple mode cannot be used in parallel.

If a ground enabled input port and a couple port have been assigned to the same input port, the ground mode and the couple mode cannot be used in parallel.

Syntax ViStatus _VI_FUNC agb220xa_couplePort(ViSession vi, ViInt16 cport_cardno, ViInt16 port1, ViInt16 port3, ViInt16 port5, ViInt16 port7, ViInt16 port9, ViInt16 port11, ViInt16 port13);

Parameters vi Instrument handle returned from agb220xa_init( ).

cport_cardno Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5 (all card) in the normal configuration mode, or 0 (all card in the auto configuration mode). For the configuration mode, see agb220xa_func.

port1 Couple port by the input ports 1 and 2. 0 (disable) or 1 (enable).

port3 Couple port by the input ports 3 and 4. 0 (disable) or 1 (enable).

port5 Couple port by the input ports 5 and 6. 0 (disable) or 1 (enable).

port7 Couple port by the input ports 7 and 8. 0 (disable) or 1 (enable).

port9 Couple port by the input ports 9 and 10. 0 (disable) or 1 (enable).

port11 Couple port by the input ports 11 and 12. 0 (disable) or 1 (enable).

VXI plug&play Driver · Driver Functions

Page 206 of the Agilent Technologies B2201A user manual — agb220xa_detectCouplePort
Page 206 · click to enlarge
Page 206

VXI plug&play Driver Driver Functions

port13 Couple port by the input ports 13 and 14. 0 (disable) or 1 (enable).

agb220xa_coupleState

This function controls the couple mode for the specified card.

Syntax ViStatus _VI_FUNC agb220xa_coupleState(ViSession vi, ViInt16 cstate_cardno, ViInt16 couple_state);

Parameters vi Instrument handle returned from agb220xa_init( ).

cstate_cardno Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5 (all card) in the normal configuration mode, or 0 (all card in the auto configuration mode). For the configuration mode, see agb220xa_func.

couple_state Couple mode. 0 (OFF) or 1 (ON).

agb220xa_dcl

This function sends a device clear (DCL) to the instrument.

A device clear will abort the present operation and enable the instrument to accept a new command or query.

This is particularly useful in situations where it is not possible to determine the instrument state. In this case, it is customary to send a device clear before issuing a new instrument driver function. The device clear ensures that the instrument will be able to begin processing the new commands.

Syntax ViStatus _VI_FUNC agb220xa_dcl(ViSession vi);

Parameters vi Instrument handle returned from agb220xa_init( ).

agb220xa_detectCouplePort

This function detects the input ports connected to the Kelvin cable, and assigns them as the input couple ports that will be used for the Kelvin connection. The input couple port setting is effective for the all cards. This command overwrites the previous couple port setting.

Syntax ViStatus _VI_FUNC agb220xa_detectCouplePort(ViSession vi);

Page 207 of the Agilent Technologies B2201A user manual — agb220xa_error_query
Page 207 · click to enlarge
Page 207

Parameters vi Instrument handle returned from agb220xa_init( ).

agb220xa_error_message

This function translates the error return value from an instrument driver function to a readable string.

Syntax ViStatus _VI_FUNC agb220xa_error_message(ViSession vi, ViStatus error_number, ViChar_VI_FAR message[ ] );

Parameters vi Instrument handle returned from agb220xa_init( ).

error_number Error return value from the driver function.

message[ ] Error message string. This is limited to 256 characters.

agb220xa_error_query

This function returns the error numbers and corresponding error messages in the error queue of a instrument. See Chapter 9 for a listing of the instrument error numbers and messages.

Instrument errors may occur when you places the instrument in a bad state such as sending an invalid sequence of coupled commands. Instrument errors can be detected by polling. Automatic polling can be accomplished by using the agb220xa_errorQueryDetect function.

Syntax ViStatus _VI_FUNC agb220xa_error_query(ViSession vi, ViPInt32 error_number, ViChar_VI_FAR error_message[ ] );

Parameters vi Instrument handle returned from agb220xa_init( ).

error_number Instrument’s error code.

error_message[ ] Instrument’s error message. This is limited to 256 characters.

VXI plug&play Driver · Driver Functions

Page 208 of the Agilent Technologies B2201A user manual — agb220xa_errorQueryDetect_Q
Page 208 · click to enlarge
Page 208

VXI plug&play Driver Driver Functions

agb220xa_errorQueryDetect

This function enables or disables automatic instrument error checking.

If automatic error checking is enabled then the driver will query the instrument for an error at the end of each function call.

Syntax ViStatus _VI_FUNC agb220xa_errorQueryDetect(ViSession vi, ViBoolean errorQueryDetect);

Parameters vi Instrument handle returned from agb220xa_init( ).

errorQueryDetect Error checking enable (VI_TRUE) or disable (VI_FALSE).

agb220xa_errorQueryDetect_Q

This function indicates if automatic instrument error detection is enabled or disabled.

Syntax ViStatus _VI_FUNC agb220xa_errorQueryDetect_Q(ViSession vi, ViPBoolean pErrDetect);

Parameters vi Instrument handle returned from agb220xa_init( ).

pErrDetect Error checking enable (VI_TRUE) or disable (VI_FALSE).

Page 209 of the Agilent Technologies B2201A user manual — 1 “ESR_OPC”
Page 209 · click to enlarge
Page 209

agb220xa_esr_Q

This function returns the contents of the ESR register. The driver returns the equivalent messages (see Parameters).

Syntax ViStatus _VI_FUNC agb220xa_esr_Q(ViSession vi, ViChar_VI_FAR errstr[ ] );

Parameters vi Instrument handle returned from agb220xa_init( ).

errstr[ ] Response from instrument.

1 “ESR_OPC”

2 “ESR_RQL”

4 “ESR_QYE_ERROR”

8 “ESR_DEVICE_DEPENDENT_ERROR”

16 “ESR_EXECUTION_ERROR”

32 “ESR_COMMAND_ERROR”

64 “ESR_URQ”

128 “ESR_PON”

OTHERS “ESR_MULTI_EVENT”

agb220xa_func

This function is used to set the channel configuration to the auto configuration mode or the normal configuration mode.

Syntax ViStatus _VI_FUNC agb220xa_func(ViSession vi, ViInt16 channel_config);

Parameters vi Instrument handle returned from agb220xa_init( ).

channel_config Configuration mode. 0 (auto) or 1 (normal).

VXI plug&play Driver · Driver Functions · Bit Value Message

Page 210 of the Agilent Technologies B2201A user manual — agb220xa_groundChanList
Page 210 · click to enlarge
Page 210

VXI plug&play Driver Driver Functions

agb220xa_groundChanCard

This function ground-enables or disables all the output ports of the specified card.

Syntax ViStatus _VI_FUNC agb220xa_groundChanCard(ViSession vi, ViInt16 disable_enable, ViInt16 gnd_cardno);

Parameters vi Instrument handle returned from agb220xa_init( ).

disable_enable Status of the card, ground enabled or ground disabled.

0. sets ground enabled card.
1. sets ground disabled card.

gnd_cardno Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5 (all card) in the normal configuration mode, or 0 (all card in the auto configuration mode). For the configuration mode, see agb220xa_func.

agb220xa_groundChanList

This function ground-enables or disables all the output ports specified by the gndchan_list.

The parameter “gndchan_list” is an array of integers with each integer representing one channel. The last number of the “gndchan_list” should be “0” (numeric zero) to identify the end of the list. The maximum number of channels that can be specified by the list is 100.

Syntax ViStatus _VI_FUNC agb220xa_groundChanList(ViSession vi, ViInt16 gndchan_disen, ViInt32_VI_FAR gndchan_list[ ] );

Parameters vi Instrument handle returned from agb220xa_init( ).

gndchan_disen Status of the port, ground enabled or ground disabled.

0. sets ground enabled port.
1. sets ground disabled port.

gndchan_list[ ] Channel numbers. 5 digits integer. ABCDE. where A: card number, BC: input port number, DE: output port number. Top zero(s) can be ignored. For example, if A=0, BC=01, and DE=01, channel number should be 101 instead of 00101.

Page 211 of the Agilent Technologies B2201A user manual — agb220xa_groundChanList_Q
Page 211 · click to enlarge
Page 211

agb220xa_groundChanList_Q

This function will query the instrument for the status, ground enabled or disabled, for the channels given in the list.

The parameter “gndchan_list” is an array of integers with each integer representing one channel. The last number of the “gndchan_list” should be “0” (numeric zero) to identify the end of the list. The maximum number of channels that can be specified by the list is 100.

The “gnd_status” parameter is an array of integers containing the return values of the query. The “gnd_status” array returned will correspond one to one with “gndchan_list” parameter.

Syntax ViStatus _VI_FUNC agb220xa_groundChanList_Q(ViSession vi, ViInt16 gnd_disen, ViInt32_VI_FAR gndchan_list[ ], ViInt32_VI_FAR gnd_status[ ] );

Parameters vi Instrument handle returned from agb220xa_init( ).

gnd_disen Query type.

0. checks if the port is the ground enabled.
1. checks if the port is the ground disabled.

gndchan_list[ ] Channel numbers to check the status. 5 digits integer. ABCDE. where A: card number, BC: input port number, DE: output port number. Top zero(s) can be ignored. For example, if A=0, BC=01, and DE=01, channel number should be 101 instead of 00101.

gnd_status[ ] Status of the channels given in the gndchan_list. Returned value depends on the setting of gnd_disen as shown below:

when gnd_disen=0, 0 means ground disabled, 1 means enabled.

when gnd_disen=1, 0 means ground enabled, 1 means disabled.

VXI plug&play Driver · Driver Functions

Page 212 of the Agilent Technologies B2201A user manual — agb220xa_groundState
Page 212 · click to enlarge
Page 212

VXI plug&play Driver Driver Functions

agb220xa_groundPort

This function will select which input port is the ground port on the specified card. For each card, you can specify the same or different ground port.

Syntax ViStatus _VI_FUNC agb220xa_groundPort(ViSession vi, ViInt16 gport_cardno, ViInt16 gnd_port);

Parameters vi Instrument handle returned from agb220xa_init( ).

gport_cardno Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5 (all card) in the normal configuration mode, or 0 (all card in the auto configuration mode). For the configuration mode, see agb220xa_func.

gnd_port Input port number to be set to the ground port. 1 to 14 (input port 1 to input port 14).

The input ground port and a ground enabled input port cannot be assigned to the same input port.

If the ground input port and a couple port have been assigned to the same input port, the ground mode and the couple mode cannot be used in parallel.

agb220xa_groundState

This function controls the ground mode for the specified card. When ground mode is ON, the input ground port is connected to all ground enabled input ports and output ports that are not connected to any other input ports. Ground disabled output ports are never connected to the input ground port when ground mode is ON.

The ground mode cannot be set to ON when the bias mode is ON.

Syntax ViStatus _VI_FUNC agb220xa_groundState(ViSession vi, ViInt16 gstate_cardno, ViInt16 state);

Parameters vi Instrument handle returned from agb220xa_init( ).

gstate_cardno Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5 (all card) in the normal configuration mode, or 0 (all card in the auto configuration mode). For the configuration mode, see agb220xa_func.

state Ground mode. 0 (OFF) or 1 (ON).

Page 213 of the Agilent Technologies B2201A user manual — agb220xa_opc_Q
Page 213 · click to enlarge
Page 213

agb220xa_init

This function initializes the software connection to the instrument and optionally verifies that instrument is in the system. In addition, it may perform any necessary actions to place the instrument in its reset state.

If the agb220xa_init function encounters an error, then the value of the vi output parameter will be VI_NULL.

Syntax ViStatus _VI_FUNC agb220xa_init(ViRsrc InstrDesc, ViBoolean id_query, ViBoolean do_reset, ViPSession vi);

Parameters InstrDesc Instrument description. Examples; GPIB0::1::INSTR.

id_query VI_TRUE (to perform In-System Verification), or VI_FALSE (do not perform In-System Verification).

do_reset VI_TRUE (to perform reset operation), or VI_FALSE (do not perform reset operation).

vi Instrument handle. This is VI_NULL if an error occurred during the initialization.

agb220xa_opc_Q

This function does the *OPC? common command.

Syntax ViStatus _VI_FUNC agb220xa_opc_Q(ViSession vi, ViPBoolean result);

Parameters vi Instrument handle returned from agb220xa_init( ).

result VI_TRUE (Operation complete), or VI_FALSE (Operation is pending).

VXI plug&play Driver · Driver Functions

Page 214 of the Agilent Technologies B2201A user manual — agb220xa_openList
Page 214 · click to enlarge
Page 214

VXI plug&play Driver Driver Functions

agb220xa_openCard

This function will disconnect all input ports from all output ports for the specified card.

If bias mode is ON, the input bias port is connected to all bias enabled output ports.

If ground mode is ON, the input ground port is connected to all ground enabled input ports and output ports.

Syntax ViStatus _VI_FUNC agb220xa_openCard(ViSession vi, ViInt16 open_cardno);

Parameters vi Instrument handle returned from agb220xa_init( ).

open_cardno Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5 (all card) in the normal configuration mode, or 0 (all card in the auto configuration mode). For the configuration mode, see agb220xa_func.

agb220xa_openList

This function will disconnect the input ports from the output ports specified by the channel list.

The parameter “openchan_list” is an array of integers with each integer representing one channel. The last number of the “openchan_list” should be “0” (numeric zero) to identify the end of the list. The maximum number of channels that can be specified by the list is 100.

Syntax ViStatus _VI_FUNC agb220xa_openList(ViSession vi, ViInt32_VI_FAR openchan_list[ ] );

Parameters vi Instrument handle returned from agb220xa_init( ).

openchan_list[ ] Channel numbers to disconnect. 5 digits integer. ABCDE. where A: card number, BC: input port number, DE: output port number. Top zero(s) can be ignored. For example, if A=0, BC=01, and DE=01, channel number should be 101 instead of 00101.

Page 215 of the Agilent Technologies B2201A user manual — agb220xa_readStatusByte_Q
Page 215 · click to enlarge
Page 215

agb220xa_openList_Q

This function will query the instrument for the channels open given in the “openchan_list”.

The parameter “openchan_list” is an array of integers with each integer representing one channel. The last number of the “openchan_list” should be “0” (numeric zero) to identify the end of the list. The maximum number of channels that can be specified by the list is 100.

The “open_status” parameter is an array of integers containing the return values of the query. The “open_status” array returned will correspond one to one with “openchan_list” parameter.

Syntax ViStatus _VI_FUNC agb220xa_openList_Q(ViSession vi, ViInt32_VI_FAR openchan_list[ ], ViInt32_VI_FAR open_status[ ] );

Parameters vi Instrument handle returned from agb220xa_init( ).

openchan_list[ ] Channel numbers to know the open status. 5 digits integer. ABCDE. where A: card number, BC: input port number, DE: output port number. Top zero(s) can be ignored. For example, if A=0, BC=01, and DE=01, channel number should be 101 instead of 00101.

open_status[ ] Status of the channels given in the openchan_list. 1 (opened) or 0 (closed).

agb220xa_readStatusByte_Q

This function returns the contents of the status byte register.

Syntax ViStatus _VI_FUNC agb220xa_readStatusByte_Q(ViSession vi, ViPInt16 statusByte);

Parameters vi Instrument handle returned from agb220xa_init( ).

statusByte The contents of the status byte are returned in this parameter.

VXI plug&play Driver · Driver Functions

Page 216 of the Agilent Technologies B2201A user manual — agb220xa_selectCompenFile
Page 216 · click to enlarge
Page 216

VXI plug&play Driver Driver Functions

agb220xa_reset

This function places the instrument in a default state. Before issuing this function, it may be necessary to send a device clear to ensure that the instrument can execute a reset. A device clear can be issued by invoking agb220xa_dcl function.

Syntax ViStatus _VI_FUNC agb220xa_reset(ViSession vi);

Parameters vi Instrument handle returned from agb220xa_init( ).

agb220xa_revision_query

This function returns the driver revision and the instrument firmware revision.

Syntax ViStatus _VI_FUNC agb220xa_revision_query(ViSession vi, ViChar_VI_FAR driver_rev[ ] , ViChar_VI_FAR instr_rev[ ] );

Parameters vi Instrument handle returned from agb220xa_init( ).

driver_rev[ ] Instrument driver revision. This is limited to 256 characters.

instr_rev[ ] Instrument firmware revision. This is limited to 256 characters.

agb220xa_selectCompenFile

This function specifies the compensation data file used by the agb220xa_compenC function. The file must contain the appropriate compensation coefficients for your measurement environment.

For obtaining the compensation coefficients for your environment and creating the compensation data file, see “Capacitance Compensation” on page 4-20.

Syntax ViStatus _VI_FUNC agb220xa_selectCompenFile(ViSession vi, ViString file_name);

Parameters vi Instrument handle returned from agb220xa_init( ).

file_name Compensation data file name. Use absolute path. If the value is NULL string, the default data is used.

Page 217 of the Agilent Technologies B2201A user manual — agb220xa_testClear
Page 217 · click to enlarge
Page 217

agb220xa_self_test

This function causes the instrument to perform a self-test and returns the result of that self-test. This is used to verify that an instrument is operating properly. A failure may indicate a potential hardware problem.

Syntax ViStatus _VI_FUNC agb220xa_self_test(ViSession vi, ViPInt16 test_result, ViChar_VI_FAR test_message[ ] );

Parameters vi Instrument handle returned from agb220xa_init( ).

test_result Numeric result from self-test operation. 0: No error.

test_message[ ] Self-test status message. This is limited to 256 characters.

agb220xa_testClear

This function clears the test result of the specified diagnostics item.

Syntax ViStatus _VI_FUNC agb220xa_testClear(ViSession vi, ViInt16 framecard_clear);

Parameters vi Instrument handle returned from agb220xa_init( ).

framecard_clear Test result to clear. 0 to 10. See below.

0. Test result of all test items
1. Card 1 relay test result
2. Card 2 relay test result
3. Card 3 relay test result
4. Card 4 relay test result
5. Relay test result of all cards
6. Front panel key test result
7. Controller test result
8. Light pen test result
9. LED matrix test result
10. Beeper test result

VXI plug&play Driver · Driver Functions

Page 218 of the Agilent Technologies B2201A user manual — agb220xa_timeOut
Page 218 · click to enlarge
Page 218

VXI plug&play Driver Driver Functions

agb220xa_testExec_Q

This function performs the diagnostics and returns the test result.

For details of each test, see “Selftest Menu” on page 3-34.

Syntax ViStatus _VI_FUNC agb220xa_testExec_Q(ViSession vi, ViInt16 framecard_exec, ViPInt16 exec_result);

Parameters vi Instrument handle returned from agb220xa_init( ).

framecard_exec Test item to perform. 1 to 10. See below.

1. Card 1 relay test
2. Card 2 relay test
3. Card 3 relay test
4. Card 4 relay test
5. Relay test of all cards
6. Front panel key test
7. Controller test
8. Light pen test
9. LED matrix test
10. Beeper test

exec_result Test result. 0: pass, 1: fail.

agb220xa_timeOut

This function sets a minimum timeout value for driver I/O transactions in milliseconds. The default timeout period is 2 seconds.

Syntax ViStatus_VI_FUNC agb220xa_timeOut(ViSession vi, ViInt32 timeOut);

Parameters vi Instrument handle returned from agb220xa_init( ).

timeOut I/O timeout value for all functions in the driver. in milliseconds. 0 to 2147483647.

Page 219 of the Agilent Technologies B2201A user manual — agb220xa_unusedPort
Page 219 · click to enlarge
Page 219

agb220xa_timeOut_Q

This function returns the timeout value for driver I/O transactions in milliseconds.

Syntax ViStatus_VI_FUNC agb220xa_timeOut_Q(ViSession vi, ViPInt32 pTimeOut);

Parameters vi Instrument handle returned from agb220xa_init( ).

pTimeOut Minimum timeout period that the driver can be set to, in milliseconds.

agb220xa_unusedPort

This function ground-enables the specified input ports of the specified card.

The parameter “unused_port” is an array of integers with each integer representing one port. The last number of the “unused_port” should be “0” (numeric zero) to identify the end of the array. The maximum number of ports that can be specified by the array is agb220xa_UNUSEDPORT_MAX.

Syntax ViStatus _VI_FUNC agb220xa_unusedPort(ViSession vi, ViInt16 unused_cardno, ViInt32_VI_FAR unused_port[ ] );

Parameters vi Instrument handle returned from agb220xa_init( ).

unused_cardno Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5 (all card) in the normal configuration mode, or 0 (all card in the auto configuration mode). For the configuration mode, see agb220xa_func.

unused_port[ ] Input ports to be ground enabled. 1 to 9 are available. Multiple port numbers can be set.

The input ground port and a ground enabled input port cannot be assigned to the same input port.

If the ground enabled input port and a couple port have been assigned to the same input port, the ground mode and the couple mode cannot be used in parallel.

VXI plug&play Driver · Driver Functions

Page 220 of the Agilent Technologies B2201A user manual — agb220xa_timeOut_Q
Page 220 · click to enlarge
Page 220

VXI plug&play Driver Driver Functions

Page 221 of the Agilent Technologies B2201A user manual — 7 Error Messages
Page 221 · click to enlarge
Page 221

7 Error Messages

Page 222 of the Agilent Technologies B2201A user manual — 7 Error Messages
Page 222 · click to enlarge
Page 222

This chapter lists and describes the error messages for Agilent B2200. An error message consists of an error number and message.

There are two types of error messages:

  • Standard SCPI Error Messages

Negative error numbers (Command Error, Execution Error, Device-Dependent Error, and Query Error) are standard SCPI errors.

  • B2200 Specific Error Messages

Positive error numbers are the B2200 specific errors.

Error messages are classified by error number as listed in the following table:

Error Range Error Category

-100 to -199 Command Error

-200 to -299 Execution Error

-300 to -399 Device-Dependent Error

-400 to -499 Query Error

100 to 32767 B2200 Specific Error

When an error occurs, the corresponding bit is set in the Standard Event Status Register (see Chapter 5):

Error Category Standard Event Status Register Bit

B2200 Specific Error bit3

Also, the error number and message are placed in the error queue, which can be read by the :SYSTem:ERRor? query command. The error queue is cleared by the common command *CLS, and when power is turned on. For these commands, see Chapter 5.

Show 6 more labels / short text on this page
Error Messages · 0 No error · Command Error bit5 · Execution Error bit4 · Device-Dependent Error bit3 · Query Error bit2
Page 223 of the Agilent Technologies B2201A user manual — Command Error
Page 223 · click to enlarge
Page 223

Error Messages Standard SCPI Error Messages

Standard SCPI Error Messages

Standard SCPI error messages have negative error numbers. The following are the standard SCPI error messages for the B2200:

Command Error

If syntax of SCPI command is not valid, a -1XX error occurs. The error number and message are placed in the error queue, and bit5 of the Standard Event Status Register is set.

A SCPI command consists of a command header and zero or more parameters. The following are example SCPI commands:

  • Subsystem command: :ROUT:OPEN:CARD 1

:ROUT:OPEN:CARD is the command header, and 1 is the parameter.

A subsystem command header consists of mnemonics (keywords) separated by colons. Query commands have a question mark (?) at end of last mnemonic.

  • Common command: *SRE 48

*SRE is the command header, and 48 is the parameter.

A common command header consists of an asterisk (*) followed by 3 characters. Query commands have a question mark (?) at end of header.

Error Number Error Message and Description

-100 Command error

Generic syntax error that cannot be determined more specifically.

-101 Invalid character

A syntax element contains a character that is invalid for that type of element; for example, a header containing an ampersand.

-102 Syntax error

An unrecognized command or data type was received; for example, a string was received when the B2200 does not accept strings.

Page 224 of the Agilent Technologies B2201A user manual — Standard SCPI Error Messages
Page 224 · click to enlarge
Page 224

Error Messages Standard SCPI Error Messages

Error Number Error Message and Description

-103 Invalid separator

An illegal character was received when a separator was expected; for example, the semicolon was omitted between multiple commands in a program message.

-104 Data type error

An improper data type was received; for example, numeric data was expected but string data was received.

-105 GET not allowed

A group execute trigger was received within a program message.

-108 Parameter not allowed

Too many parameters for the command were received.

-109 Missing parameter

Fewer parameters were received than required for the command.

-110 Command header error

An error was detected in the header. This error message is reported if the B2200 cannot determine the more specific header errors -111 through -114.

-111 Header separator error

An illegal character for a header separator was received; for example, no white space between the command header and parameter.

-112 Program mnemonic too long

A keyword in the command header contains more than twelve characters.

-113 Undefined header

An undefined command header was received; for example, *XYZ or :ROUT:OPEN:COOD.

Page 225 of the Agilent Technologies B2201A user manual — Standard SCPI Error Messages
Page 225 · click to enlarge
Page 225

Error Messages Standard SCPI Error Messages

Error Number Error Message and Description

-120 Numeric data error

An error was detected in a numeric parameter (including the non-decimal numeric types). This error message is reported when the B2200 cannot determine the more specific errors -121 through -128.

-121 Invalid character in number

An invalid character for the parameter was received; for example, an alphacharacter was received when the parameter type was decimal numeric.

-123 Exponent too large

The magnitude of the exponent for a numeric parameter was larger than 32000.

-124 Too many digits

The mantissa of a decimal numeric parameter contained more than 255 digits excluding leading zeros.

-128 Numeric data not allowed

Numeric data is not allowed in this position for this command.

-138 Suffix not allowed

A suffix was received after a numeric parameter. For the B2200, no parameters have suffix.

-140 Character data error

An error was detected in a character parameter. This error message is reported if the B2200 cannot determine the more specific errors -141 through -148.

-141 Invalid character data

Either the character parameter contains an invalid character or the particular element received is not valid for the command.

-144 Character data too long

The character parameter contains more than 12 characters.

Page 226 of the Agilent Technologies B2201A user manual — Standard SCPI Error Messages
Page 226 · click to enlarge
Page 226

Error Messages Standard SCPI Error Messages

Error Number Error Message and Description

-148 Character data not allowed

A character parameter is not allowed for this position.

-150 String data error

An error was detected in a string parameter. This error is reported if the B2200 cannot determine a more specific error -151 and -158.

-151 Invalid string data

An invalid string parameter data was received; for example, an END message was received before the terminal quote character.

-158 String data not allowed

A string parameter data was received but was not allowed at this point.

-160 Block data error

An error was detected in a block data. This error is reported if the B2200 cannot determine more specific errors -161 and -168.

-161 Invalid block data

An invalid block data was received; for example, an END message was received before the length was satisfied.

-168 Block data not allowed

A legal block data was received but was not allowed at this point.

-170 Expression error

An error was detected in an expression. This error is reported if the B2200 cannot determine more specific errors -171 and -178.

-171 Invalid expression

The expression was invalid; for example, unmatched parentheses or an illegal character.

-178 Expression data not allowed

An expression was received but was not allowed at this point.

Page 227 of the Agilent Technologies B2201A user manual — Execution Error
Page 227 · click to enlarge
Page 227

Error Messages Standard SCPI Error Messages

Execution Error

If syntax of a SCPI command header and parameter is valid, but the command cannot be executed due to some condition of the B2200, a -2XX error occurs. The error number and message are placed in the error queue, and bit4 of the Standard Event Status Register is set.

Error Number Error Message and Description

-200 Execution error

Generic execution error that cannot be determined more specifically.

-220 Parameter error

The parameter value is valid, but not executable due to some condition of the B2200. This error occurs if more specific errors -221 through -224 cannot be determined.

-222 Data out of range

A valid parameter setting could not be executed because interpreted value was out of range as defined by the B2200.

-223 Too much data

A valid parameter setting (block, expression, or string type) could not be executed because parameter contained more data than the B2200 could handle due to insufficient memory or other requirements.

-224 Illegal parameter value

A valid parameter setting (where exact value from list of possibles was expected) could not be executed due to present B2200 state.

-260 Expression error

An expression related error occurred.

Page 228 of the Agilent Technologies B2201A user manual — Device-Dependent Errors
Page 228 · click to enlarge
Page 228

Error Messages Standard SCPI Error Messages

Device-Dependent Errors

-3XX errors indicate that an B2200 operation did not properly complete, possibly due to an abnormal hardware or firmware condition. These negative codes are SCPI defined. For the device-dependent positive error codes, see “B2200 Specific Error Messages” on page 7-10. The positive codes are not SCPI defined.

For these errors, an error number and message are placed in the error queue, and bit3 of the Standard Event Status Register is set.

Error Number Error Message and Description

-300 Device-specific error

Generic device-dependent error for the B2200 that cannot be determined more specifically.

-311 Memory error

An error was detected in the B2200's memory.

-350 Queue overflow

If error queue is full, this code is entered into the error queue instead of the code that caused the error. This code indicates that there was no room in the error queue, so an error occurred but was not recorded.

Page 229 of the Agilent Technologies B2201A user manual — Query Errors
Page 229 · click to enlarge
Page 229

Error Messages Standard SCPI Error Messages

Query Errors

If the output queue control of the B2200 detects one of following problems, a -4XX error occurs:.

  • An attempt was made to read data from the output queue when no output data is present or pending.
  • Data in the output queue has been lost.

If this type of error occurs, the error number and message are placed in the error queue, and bit2 of the Standard Event Status Register is set.

Error Number Error Message and Description

-400 Query error

Generic query error for the B2200 that cannot be determined more specifically.

-410 Query INTERRUPTED

A condition causing an INTERRUPTED query error occurred; for example, a query followed by DAB or GET before a response was completely sent.

-420 Query UNTERMINATED

A condition causing an UNTERMINATED query error occurred; for example, the B2200 was addressed to talk and an incomplete program message was received.

-430 Query DEADLOCKED

A condition causing a DEADLOCKED query error occurred; for example, both input buffer and output buffer are full and the B2200 cannot continue.

-440 Query UNTERMINATED after indefinite response

A query was received in the same program message after a query requesting an indefinite length response was executed.

Page 230 of the Agilent Technologies B2201A user manual — B2200 Specific Error Messages
Page 230 · click to enlarge
Page 230

Error Messages B2200 Specific Error Messages

B2200 Specific Error Messages

These are the B2200-specific errors that are not defined by SCPI. These errors indicate that an B2200 operation did not properly complete due to card, channel, port, or mode errors.

For the SCPI defined device-dependent codes, see “Device-Dependent Errors” on page 7-8.

For these errors, an error number and message are placed in the error queue, and bit3 of the Standard Event Status Register is set.

The B2200 specific error messages have positive error numbers.

Page 231 of the Agilent Technologies B2201A user manual — B2200 Channel Related Errors
Page 231 · click to enlarge
Page 231

Error Messages B2200 Specific Error Messages

B2200 Channel Related Errors

Error Number Error Message and Description

2000 Invalid card number

Wrong card number is specified in card number or channel list parameter. Make sure card is properly installed in the correct B2200 slot.

2001 Invalid channel number

Wrong channel number is specified in the channel list. Confirm the channel number, the card configuration, and the configuration mode of the B2200.

2006 Command not supported on this card

Command was used that is not supported by the card.

2009 Too many channels in channel list

Too many channels are specified in the channel list (maximum 120).

2011 Empty channel list

No channel list is specified.

2012 Invalid channel range

Wrong channel list is defined. Confirm the syntax of the channel list, the card configuration, and the configuration mode of the B2200.

2002 Unsupported Module · 2003 Unsupported Configuration

Page 232 of the Agilent Technologies B2201A user manual — B2200 Card/Mode/Port Related Errors
Page 232 · click to enlarge
Page 232

Error Messages B2200 Specific Error Messages

B2200 Card/Mode/Port Related Errors

Error Number Error Message and Description

3000 Card0 initialization fail

The B2200 may be defective. Contact your nearest Agilent Technologies service center.

3001 Card1 initialization fail

The B2200 or card installed in slot 1 of the B2200 may be defective.

3002 Card2 initialization fail

The B2200 or card installed in slot 2 of the B2200 may be defective.

3003 Card3 initialization fail

The B2200 or card installed in slot 3 of the B2200 may be defective.

3004 Card4 initialization fail

The B2200 or card installed in slot 4 of the B2200 may be defective.

3011 Bad couple port number

For the Input Port No. for the couple port, only 1, 3, 5, 7, or 9 are allowed.

3012 Bad bias port number

For Input Port No. for Bias Port, only 1 to 10 are allowed.

3013 Cannot connect multiple channels in SROUTe mode

For the single connection rule, an input port can be connected to only one output port, and an output port can be connected to only one input port.

3014 Cannot directly specify Bias Port channel

Relays on the input bias port cannot be controlled directly when the bias mode is ON.

Page 233 of the Agilent Technologies B2201A user manual — B2200 Card/Mode/Port Related Errors
Page 233 · click to enlarge
Page 233

Error Messages B2200 Specific Error Messages

Error Number Error Message and Description

3017 Too many relays closed. Max 52 relays/card.

Too many relays have been closed. To close new relays, open some relays. Maximum 52 relays can be closed for each module.

3018 Can't change to ACONfig mode. Check card config.

The present card configuration of the B2200 does not allow the auto configuration mode. Change the module configuration.

3019 Cannot use same port for Couple and Bias

The couple mode and the bias mode cannot be used in parallel when a couple port and the input bias port have been assigned to the same input port.

3020 Bad auto ground port number

Specify the output ports effective for the input ground port. The port number must be 1 to 14

3021 Bad unused port number

Specify the input ports effective for the ground enabled input port (unused port). The port number must be 1 to 9.

3022 Cannot directly specify auto ground port channel

Relays on the input ground port cannot be controlled directly when the ground mode is ON.

3023 Cannot directly specify unused port channel

Relays on the ground enabled input port (unused port) cannot be controlled directly when the ground mode is ON.

3024 Cannot use same port for Couple and Auto Ground

The couple mode and the ground mode cannot be used in parallel when a couple port and the input ground port have been assigned to the same input port.

3025 Cannot use same port for Unused and Auto Ground

The input ground port and the ground enabled input port (unused port) cannot be assigned to the same input port.

Page 234 of the Agilent Technologies B2201A user manual — B2200 Card/Mode/Port Related Errors
Page 234 · click to enlarge
Page 234

Error Messages B2200 Specific Error Messages

Error Number Error Message and Description

3026 Cannot use same port for Unused and Couple

The couple mode and the ground mode cannot be used in parallel when a couple port and the ground enabled input port (unused port) have been assigned to the same input port.

3027 Cannot use Unused Port during Auto Ground Mode ON

The ground enabled input port (unused port) must be opened during the ground mode is ON.

3030 Bad input port number

The input port number must be 1 to 14.

3031 Bad output port number

The output port number must be 1 to 12, 24, 36, or 48. Maximum number depends on the switch module configuration.

3032 Bad setting memory number

The internal memory number for the setup data must be 1 to 8.

3033 EEPROM programming failure

Cannot change the GPIB address or update internal memory data. Contact your nearest Agilent Technologies service center.

3034 EEPROM reading failure

Cannot read the GPIB address or internal memory data. Contact your nearest Agilent Technologies service center.

3035 Setting memory data is invalid

Cannot read the setup data in the internal memory. The data has been broken. Delete the data.

3036 Cannot load this setting data in this configuration

Cannot read the setup data in the internal memory due to the configuration mismatch. Delete the data or leave it until restoring the module configuration.

Display codes

Error codes

What the fault codes shown on the display mean and how to clear them.

CodeMeaningFix
-300Device-specific error Generic device-dependent error for the B2200 that cannot be determined more specifically. p.228
-311Memory error An error was detected in the B2200's memory. p.228
-350Queue overflow If error queue is full, this code is entered into the error queue instead of the code that caused the error. This code indicates that there was no room in the error queue, so an error occurred but was not recorded. p.228
-400Query error Generic query error for the B2200 that cannot be determined more specifically. p.229
-410Query INTERRUPTED A condition causing an INTERRUPTED query error occurred; for example, a query followed by DAB or GET before a response was completely sent. p.229
-420Query UNTERMINATED A condition causing an UNTERMINATED query error occurred; for example, the B2200 was addressed to talk and an incomplete program message was received. p.229
-430Query DEADLOCKED A condition causing a DEADLOCKED query error occurred; for example, both input buffer and output buffer are full and the B2200 cannot continue. p.229
-440Query UNTERMINATED after indefinite response A query was received in the same program message after a query requesting an indefinite length response was executed. p.229

Ask about the Agilent Technologies B2201A

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Agilent Technologies B2201A

Be the first to ask — other owners or our team will answer.

Original document

Agilent Technologies B2201A manual PDF

234 pages·1.2 MB·Switch·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…