Agilent B2200A Femto Leakage Switch Mainframe Agilent B2201A 14ch Low Leakage Switch Mainframe

© Agilent Technologies 2004, 2005
defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
No part of this manual may be reproduced in any form or by any means (including elec- tronic storage and retrieval or translation into a foreign language) without prior agree- ment and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.
The material contained in this docu- ment is provided “as is,” and is sub- ject to being changed, without notice, in future editions. Further, to the max- imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par- ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connec- tion with the furnishing, use, or per- formance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep- arate agreement shall control.
B2200-90000
Edition
Edition 1, October 2004 Edition 2, July 2005
Agilent Technologies, Inc. 5301 Stevens Creek Blvd Santa Clara, CA 95051 USA
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor- dance with the terms of such license.
If software is for use in the performance of a U.S. Government prime contract or subcon- tract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer soft- ware” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclo- sure of Software is subject to Agilent Tech- nologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as

DECLARATION OF CONFORMITY According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014
Rue de la Gare 29 CH - 1110 Morges Switzerland
Declares under sole responsibility that the product as originally delivered
Femto Leakage Switch Module 14ch Low Leakage Switch Mainframe 14ch Low Leakage Switch Module Model Number: Agilent B2200A
Agilent B2210A Agilent B2201A
complies with the essential requirements of the following applicable European Directives, and carries the CE marking accordingly:
Low Voltage Directive (73/23/EEC, amended by 93/68/EEC) EMC Directive (89/336/EEC, amended by 93/68/EEC) and conforms with the following product standards EMC Standard IEC 61326:2002 / EN 61326:1997 +A1:1998 +A2:2001 +A3:2003 CISPR 11:1997 / EN 55011:1998
Group 1 Class A 4 kV CD, 8 kV AD 3 V/m, 80-1000 MHz 0.5 kV signal lines, 1 kV power lines 0.5 kV line-line, 1 kV line-ground 3 V, 0.15-80 MHz 1 cycle, 100%
IEC61000-4-2:1995 / EN61000-4-2:1995 IEC 61000-4-3:1995 / EN61000-4-3:1995 IEC 61000-4-4:1995 / EN61000-4-4:1995 IEC 61000-4-5:1995 / EN61000-4-5:1995 IEC 61000-4-6:1996 / EN61000-4-6:1996 IEC 61000-4-11:1994 / EN61000-4-11:1994
The product was tested in a typical configuration with Agilent Technologies test systems.
Safety IEC 61010-1:2001 / EN 61010-1:2001 Canada: CSA C22.2 No. 1010.1:1992, NRTL/C
QA Manager Agilent Technologies
Limit · September 26, 2006 · Date Toshiyuki Kawaji

Lpa < 70 dB
nach DIN 45635 T. 19
Lpa < 70 dB
Cet appareil ISM est conforme ?Hla norme NMB-001 du Canada.
This product complies with the WEEE Directive (2002/96/EC) marking requirements. The affixed label indicates that you must not discard this electrical/ electronic product in domestic household waste.
Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as a “Monitoring and Control instrumentation” product.
Do not dispose in domestic household waste.
To return unwanted products, contact your local Agilent office, or see www.agilent.com/environment/product/ for more information.
Microsoft, Windows, Windows NT, Visual C++, and Visual Basic are registered trademarks of Microsoft Corporation.
Borland C++ Builder is registered trademark of International, Inc. LabWindows and LabVIEW are registered
trademarks of National Instruments Corporation. All other trademarks are the property of their respective owners.

The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual may impair the protections provided by the equipment. In addition, it violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies, Inc. assumes no liability for customer’s failure to comply with these requirements.
Agilent B2200 is INDOOR USE products.
This is Safety Class I instrument. To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical ground. The power terminal and the power cable must meet International Electrotechnical Commission (IEC) safety standards.
Do not operate the instrument in the presence of flammable gases or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard.
Operation personnel must not remove instrument covers. Component replacement and internal adjustments must be made by qualified maintenance personnel. Do not replace components with power cable connected. Under certain conditions, dangerous voltages may exist even with the power cable removed. To avoid injuries, always disconnect power and discharge circuits before touching them.
Do not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present.

Because of the danger of introducing additional hazards, do not install substitute parts or perform any unauthorized modification to the instrument. Return the instrument to a Agilent Technologies Sales and Service Office for services and repair to ensure that safety features are maintained.

The general definitions of safety symbols used on equipment or in manuals are listed below.
Instruction manual symbol: the product will be marked with this symbol when it is necessary for the user to refer to the instruction manual in order to protect against damage to the instrument.
Indicates dangerous voltage and potential for electrical shock. Do not touch terminals that have this symbol when insrument is on.
Affixed to product containing static sensitive devices--use anti-static handling procedures to prevent electrostatic discharge damage to component.
Protective conductor terminal. For protection against electrical shock in case of a fault. Used with field wiring terminals to indicate the terminal which must be connected to ground before operating equipment.
Frame or chassis terminal. A connection to the frame (chassis) of the equipment which normally includes all exposed metal structures.
Indicates earth (ground) terminal.
Alternating current.
Direct current.
ON (Supply).
OFF (Supply).
STANDBY (Supply).
Means INSTALLATION CATEGORY I. Measurement terminals on the rear panel comply with INSTALLATION CATEGORY I.

This manual is a user’s guide for Agilent B2200A/B2201A Switch Mainframe, and consists of the following chapters:
Describes an overview and specifications of the Agilent B2200 series.
Describes how to install the Agilent B2200 and how to setup the input/output.
Explains the front panel operation and the switch control functions of the Agilent B2200, also provides the reference information of the front panel keys and display.
Explains the programming examples that control the Agilent B2200.
Describes the all commands of the Agilent B2200 SCPI commands.
Describes the all functions of the Agilent B2200 VXIplug&play driver.
Lists and describes the error messages for the Agilent B2200.
The following text conventions are used in this manual:
Screen Text Represents text that appears on screen of the controller.
Italic Refers to a related document, or is used for emphasis.
Text Conventions

Agilent B2200 Series 1-3
Front Panel 1-4
1-4 Rear Panel 1-6
Switch Modules 1-8
1-8
Specifications 1-10
General Specifications 1-11 Switch Modules 1-12 Supplemental Information for B2200A/B2210A 1-13 Supplemental Information for B2201A/B2211A 1-14
Accessories and Options 1-15
2-2
Requirements 2-3
Power Requirements 2-3 Power Cable 2-3 Operating Environment 2-5 Storage and Shipping Environment 2-5
Inspection 2-6
Installing the B2200 2-7
To Set the GPIB Address 2-8 To Connect the GPIB Cable 2-8 To Install the Switch Module 2-9
2-9 To Install the Blank Panel 2-9

Self-Test 2-10
Output Connections 2-11
Output Connectors 2-11 Connector Plates 2-12 To Make Connections to DUT Interface 2-13 To Make Interlock Circuit 2-15 To Mount Connectors 2-18
Input Connections 2-20
Measurement Cable Length 2-23
Maintenance 2-25
Calibration 2-25 Cleaning 2-25
Operation 3-3
To Initialize Agilent B2200 3-4 To Enable Light Pen 3-4 To Change Channel Configuration Mode 3-4 To Change Connection Rule 3-5 To Change Connection Sequence 3-5 To Control Switch Condition 3-5 To Open All Switches 3-6 To Save/Load Setup Data 3-6 To Use Bias Mode 3-7 To Use Ground Mode 3-8 To Use Couple Mode 3-9 To Display Firmware Revision 3-9 To Display Module Information 3-9 To Read Error Message 3-10 To Set Beeper 3-10

To Set GPIB Address 3-10 To Set Remote Display Mode 3-11 To Return to Local Mode 3-11
Switch Control Functions 3-12
Channel Configuration Mode 3-13 Connection Rule 3-14 Connection Sequence 3-15 Bias Mode 3-16 Ground Mode 3-18 Couple Mode 3-20
Display Functions 3-22
LED Matrix 3-22 LCD 3-23
Front Panel Keys 3-25
Function key group 3-26 Edit key group 3-29
Setup Menus 3-30
Setup Menu 3-30 Selftest Menu 3-34
Programming Basics 4-3
SCPI Command Hierarchy 4-3 Fundamental Commands 4-4 Switch Control 4-5
Programming Examples 4-7
Connecting Input-Output Paths 4-8 Using Bias Mode 4-10 Using Ground Mode 4-12

Using Couple Mode 4-14 Saving Input/Output Labels 4-16 Defining Comment for Internal Memory 4-18
Capacitance Compensation 4-20
Capacitance Compensation Function 4-20 Required Conditions 4-21 To Create Compensation Data File 4-23 To Perform Measurement and Compensation 4-27
Common Commands 5-4
Commands Summary 5-4 *CLS 5-5 *ESE 5-5 *ESR? 5-6 *IDN? 5-7 *OPC 5-8 *RST 5-9 *SRE 5-10 *STB? 5-11 *TST? 5-12 *WAI 5-12
Subsystem Commands 5-13
Command Summary 5-13 :DIAGnostic:TEST:CARD:CLEar 5-21 :DIAGnostic:TEST:CARD[:EXECute]? 5-21 :DIAGnostic:TEST:CARD:STATe? 5-22 :DIAGnostic:TEST:FRAMe:CLEar 5-22 :DIAGnostic:TEST:FRAMe[:EXECute]? 5-23 :DIAGnostic:TEST:FRAMe:STATe? 5-24 [:ROUTe]:AGND:CHANnel:DISable:CARD 5-24

[:ROUTe]:AGND:CHANnel:DISable[:LIST] 5-25 [:ROUTe]:AGND:CHANnel:ENABle:CARD 5-25 [:ROUTe]:AGND:CHANnel:ENABle[:LIST] 5-26 [:ROUTe]:AGND:PORT 5-27 [:ROUTe]:AGND[:STATe] 5-28 [:ROUTe]:AGND:UNUSED 5-29 [:ROUTe]:BIAS:CHANnel:DISable:CARD 5-30 [:ROUTe]:BIAS:CHANnel:DISable[:LIST] 5-30 [:ROUTe]:BIAS:CHANnel:ENABle:CARD 5-31 [:ROUTe]:BIAS:CHANnel:ENABle[:LIST] 5-31 [:ROUTe]:BIAS:PORT 5-32 [:ROUTe]:BIAS[:STATe] 5-33 [:ROUTe]:CLOSe:CARD? 5-34 [:ROUTe]:CLOSe[:LIST] 5-34 [:ROUTe]:CONNection:RULE 5-35 [:ROUTe]:CONNection:SEQuence 5-36 [:ROUTe]:COUPle:PORT 5-37 [:ROUTe]:COUPle:PORT:DETect 5-38 [:ROUTe]:COUPle[:STATe] 5-39 [:ROUTe]:FUNCtion 5-40 [:ROUTe]:OPEN:CARD 5-40 [:ROUTe]:OPEN[:LIST] 5-41 [:ROUTe]:SYMBol:CHANnel 5-42 [:ROUTe]:SYMBol:PORT 5-43 :SYSTem:BEEP 5-44 :SYSTem:CCONfig? 5-44 :SYSTem:CDEScription? 5-45 :SYSTem:CPON 5-46 :SYSTem:CTYPe? 5-47 :SYSTem:DISPlay:LCD 5-47 :SYSTem:DISPlay:LED 5-48 :SYSTem:DISPlay:STRing 5-48

:SYSTem:ERRor? 5-49 :SYSTem:KLC 5-49 :SYSTem:MEMOry:COMMent 5-50 :SYSTem:MEMOry:DELete 5-50 :SYSTem:MEMOry:LOAD 5-51 :SYSTem:MEMOry:SAVE 5-51 :SYSTem:PEN 5-52 :SYSTem:VERSion? 5-52
Status Reporting Structure 5-53
Status Reporting Structure 5-53 Status Byte Register 5-55 Service Request Enable Register 5-57 Standard Event Status Register 5-58 Standard Event Status Enable Register 5-59 Output Queue 5-60
System Requirements 6-3
Installing VXIplug&play Driver 6-4
Driver Functions 6-5
agb220xa_biasChanCard 6-7 agb220xa_biasChanList 6-7 agb220xa_biasChanList_Q 6-8 agb220xa_biasPort 6-9 agb220xa_biasState 6-9 agb220xa_close 6-10 agb220xa_closeCard_Q 6-10 agb220xa_closeList 6-11 agb220xa_closeList_Q 6-12 agb220xa_cmd 6-12

agb220xa_cmdData_Q 6-13 agb220xa_cmdInt 6-13 agb220xa_cmdInt16Arr_Q 6-14 agb220xa_cmdInt16_Q 6-14 agb220xa_cmdInt32Arr_Q 6-15 agb220xa_cmdInt32_Q 6-15 agb220xa_cmdReal 6-16 agb220xa_cmdReal64Arr_Q 6-16 agb220xa_cmdReal64_Q 6-17 agb220xa_cmdString_Q 6-17 agb220xa_compenC 6-18 agb220xa_connRuleSeq 6-18 agb220xa_couplePort 6-19 agb220xa_coupleState 6-20 agb220xa_dcl 6-20 agb220xa_detectCouplePort 6-20 agb220xa_error_message 6-21 agb220xa_error_query 6-21 agb220xa_errorQueryDetect 6-22 agb220xa_errorQueryDetect_Q 6-22 agb220xa_esr_Q 6-23 agb220xa_func 6-23 agb220xa_groundChanCard 6-24 agb220xa_groundChanList 6-24 agb220xa_groundChanList_Q 6-25 agb220xa_groundPort 6-26 agb220xa_groundState 6-26 agb220xa_init 6-27 agb220xa_opc_Q 6-27 agb220xa_openCard 6-28 agb220xa_openList 6-28 agb220xa_openList_Q 6-29

agb220xa_readStatusByte_Q 6-29 agb220xa_reset 6-30 agb220xa_revision_query 6-30 agb220xa_selectCompenFile 6-30 agb220xa_self_test 6-31 agb220xa_testClear 6-31 agb220xa_testExec_Q 6-32 agb220xa_timeOut 6-32 agb220xa_timeOut_Q 6-33 agb220xa_unusedPort 6-33
Standard SCPI Error Messages 7-3
Command Error 7-3 Execution Error 7-7 Device-Dependent Errors 7-8 Query Errors 7-9
B2200 Specific Error Messages 7-10
B2200 Channel Related Errors 7-11 B2200 Card/Mode/Port Related Errors 7-12


This chapter describes the basic functions and features of the Agilent B2200A/B2201A (Agilent B2200 series), and consists of the following sections:
Introduction

Introduction Agilent B2200 Series
Agilent B2200 series is a switching matrix designed for semiconductor dc characteristics measurement applications. The Agilent B2200 series has 14 input ports and four card slots for the switch modules (plug-in cards), and can configure a 12, 24, 36, or 48 outputs switching matrix. The Agilent B2200A/B2201A supports the following dedicated switch module.
Dedicated for the Agilent B2200A mainframe.
Dedicated for the Agilent B2201A mainframe.
Basic functions of the Agilent B2200 series are listed below.

The Agilent B2200 series provides the front panel keys, the LCD, and the LED matrix display for the status monitor and connection setup.
Used to turn the Agilent B2200 on or off.
Used to monitor the status and set the connection. See Chapter 3 for the details.
Used to set the Agilent B2200. See Chapter 3 for the details.
Displays the status of the matrix switches. Also used to set the matrix connection with the light pen.
Inputs for the source monitor unit (SMU). Eight input ports. Up to four kelvin inputs are available. Triaxial BNC connector.
Multipurpose inputs. Six input ports. Coaxial BNC connector. CMH and CML terminals are the input ports for the capacitance measurement.
The input port 12 (AUX Input 12) is the default ground input port. If you assign the port as the ground input port, this ground port will be internally connected to the ground when the ground mode is set to ON. In this case, open this connector.
Connector for the Agilent 16443A Light pen. Used to set the matrix connection with the LED matrix display.
Do not apply an input signal over these limits to the input terminals. If you do, the Agilent B2200 will be damaged. If you use a bias source that has current limit capability, set the bias source current limit.
Introduction Front Panel

Figure 1-1 Front Panel View

Introduction · Front Panel

The Agilent B2200 series has four card slots for the switch modules, the GPIB interface and so on.
For the Agilent B2200A mainframe, the Agilent B2210A cards are installed.
For the Agilent B2201A mainframe, the Agilent B2211A cards are installed.
Mixed configuration of the switch modules is not supported.
Use an Agilent 82357A USB/GPIB interface or Agilent 10833A/B/C/D GPIB cable to connect to an external computer or equipment.
You need this serial number when using the Agilent Technologies telephone assistance program.
AC power cable is connected to this receptacle.
Introduction Rear Panel

Figure 1-2 Rear Panel View

Introduction · Rear Panel


The Agilent B2200A and B2201A support the dedicated switch module, Agilent B2210A and B2211A, respectively.
By installing the modules, the module inputs will be connected internally to the front panel input connectors. And 12 output connectors will face the rear panel. The type of the output connectors is the triaxial BNC. Up to six kelvin outputs are available.
Image of the output terminals and the block diagram of the switch module are shown in Figure 1-3 and Figure 1-4.
Figure 1-3 Switch Module Output Terminals


Figure 1-4 Switch Module Block Diagram
1 2 3 4 5 6 7 8 9 10 11 12
14 CML
Introduction Switch Modules · Output · Input

This section lists specifications and supplemental information for the Agilent B2200 series.
The specifications are the performance standards or limits against which these units have been tested. The supplemental information is not warranted, but provides useful information about functions and performance.
If not noted otherwise, the conditions for specifications and supplemental information are as follows:
Humidity: 5 % to 60 % R.H.
Introduction Specifications

Temperature range:
Storage:
< 80 % R.H. at 35 °C, < 60 % R.H. at 65 °C, non-condensing (B2200A/B2210A)
< 80 % R.H. at 65 °C, non-condensing (B2201A/B2211A)
Regulatory compliance:
Power requirement:
90 to 264 V (continuous), 47 to 63 Hz, 2 A/200 VA maximum
Introduction Specifications

Table 1-1 lists the specifications when the switch module is installed in the mainframe. In the table, the conditions are as follows:
Conditions: 23 °C ± 5 °C, 5 % to 60 % R.H.
Table 1-1 Agilent B2210A/B2211A Switch Module Specifications
Max. current rating (A)
SMU input 1.0 1.0
AUX input 0.5 0.5
Max. voltage rating (V)
SMU input, channel to common 200 200
SMU input, channel to channel 300 300
AUX input, channel to common 100 100
AUX input, channel to channel 100 100
Closed channel residual resistance (Ω)
SMU input 0.6 0.6
AUX input 1.5 1.5
Channel isolation resistance (Ω)
SMU input 1014 5 × 1013
AUX input 109 109
The Agilent B2200A mainframe supports the Agilent B2210A module.
The Agilent B2201A mainframe supports the Agilent B2211A module.
Mixed configuration of the switch modules is not supported.
Introduction Specifications

Offset current 1 < 10 fA (typical 3 fA) (SMU input)
IM noise (RMS) 2 0.6 fA (SMU input)
Channel crosstalk capacitance < 1 pF/channel (SMU input)
< 3 pF/channel (AUX input)
Offset voltage < 50 µV (SMU input)
< 80 µV (AUX input)
Settling time 3 2.0 seconds
Bandwidth (at -3dB) 30 MHz (SMU input)
Guard capacitance 4 < 145 pF (SMU input)
Additional C measurement error 5 < ± 1 % + 0.2 pF (SMU input)
1. When the voltage applied to all input-output channels is 0 V. 2. Measured by the Agilent 4156C with the integration time setting 100 PLC. When the voltage applied to all other input-output paths is 0 V. 3. The time until the measurement value settles to within 50 fA of the final value. When the applied voltage is 10 V. 4. When four modules have been installed in the mainframe. And when only one input-output path is made. 5. For the measurement less than 1000 pF at the frequency 1 kHz to 1 MHz, using the Agilent 4284A with 3 m cable. After the compensation using the capacitance compensation algorithm (a function of VXIplug&play driver).
Introduction Specifications

Offset current 1 < 50 fA (SMU input)
IM noise (RMS) 2 5 fA (SMU input)
Channel crosstalk capacitance < 0.5 pF/channel (SMU input)
< 3 pF/channel (AUX input)
Offset voltage < 80 µV (SMU input)
< 100 µV (AUX input)
Settling time 3 2.0 seconds
Bandwidth (at -3dB) 30 MHz (SMU input)
Guard capacitance 4 < 145 pF (SMU input)
Additional C measurement error 5 < ± 1 % + 0.2 pF (SMU input)
1. When the voltage applied to all input-output paths is 0 V. 2. Measured by the Agilent 4156C with the integration time setting 100 PLC. When the voltage applied to all other input-output paths is 0 V. 3. The time until the measurement value settles to within 300 fA of the final value. When the applied voltage is 10 V. 4. When four modules have been installed in the mainframe. And when only one input-output path is made. 5. For the measurement less than 1000 pF at the frequency 1 kHz to 1 MHz, using the Agilent 4284A with 3 m cable. After the compensation using the capacitance compensation algorithm (a function of VXIplug&play driver).
Introduction Specifications

Introduction Accessories and Options
Agilent B2200 is furnished with the following accessories.
Stores the Agilent B2200 VXIplug&play driver.
The Agilent B2200 VXIplug&play driver supports Windows XP Professional, Windows 2000, Windows NT 4.0, Windows 98, and Windows 95.
Table 1-2 lists the options and accessories available for the Agilent B2200.

Introduction Accessories and Options
Table 1-2 Options and Accessories
Model Number Option Item Description
B2200A fA Leakage Switch Mainframe
B2200A-UK6 Commercial cal. certificate w/ test data
B2200A-ABA Manual set, English
B2200A-ABJ Manual set, Japanese
B2201A 14ch Low Leakage Switch Mainframe
B2201A-UK6 Commercial cal. certificate w/ test data
B2201A-ABA Manual set, English
B2201A-ABJ Manual set, Japanese
B2210A fA Leakage Switch Module (for B2200A)
B2211A 14ch Low Leakage Switch Module (for B2201A)
16493H GNDU cable (between 41501/4142 and 16495F/G)
16493H-001 1.5 m length
16493H-002 3 m length
16493J Interlock cable (between E5260/E5270/4155/4156 and 16495F/G)
16493J-001 1.5 m length
16493J-002 3 m length
16493J-003 5 m length
16493K Kelvin triaxial cable (between B2200 inputs and E5260/E5270/4156/41501)
16493K-001 1.5 m length
16493K-002 3 m length
16493L GNDU cable (between E5260/E5270 and 16495F/G)
16493L-001 1.5 m length
16493L-002 3 m length
16493L-003 5 m length
16493N GNDU cable for Kelvin connection (between B2200 inputs and E5260/E5270/41501/4142)
16443A Light pen

Introduction Accessories and Options
Model Number Option Item Description
16494A-001 1.5 m length
16494A-002 3 m length
16494A-003 80 cm length
16494A-005 4 m length
16494B Kelvin triaxial cable (between B2200 inputs and 4142B, between B2210/B2211 outputs and 16495F/G)
16494B-001 1.5 m length
16494B-002 3 m length
16494B-003 80 cm length
16494C Kelvin triaxial cable (between B2210/B2211 outputs and B2220A)
16494C-001 1.5 m length
16494C-002 3 m length
16494C-005 4 m length
16494F CMU Input cable, 2 m (between B2200 CMH/CML and 4-terminal pairs connectors)
16495E Half size blank plate
16495F Half size connector plate with 12×triaxial, intlk, GNDU
16495F-001 Bulkhead feedthrough connectors (female to female)
16495F-002 Connectors to contacts for soldering
16495G Full size connector plate with 24×triaxial, intlk, GNDU
16495G-001 Bulkhead feedthrough connectors (female to female)
16495G-002 Connectors to contacts for soldering
16494A Triaxial cable

Introduction Accessories and Options


This chapter describes requirements to install Agilent B2200 and the tasks for installation, and is organized into the following three sections:
Installation

This section describes the following requirements for the Agilent B2200.
The Agilent B2200 can operate from any single-phase ac power source supplying 90 to 264 V in the frequency range from 47 to 63 Hz. The maximum power consumption is 2 A/200 VA.
In accordance with international safety standards, this instrument is equipped with a three-wire power cable. When connected to an appropriate ac power outlet, this cable grounds the instrument frame. The type of power cable shipped with each instrument depends on the country of destination. See the following table for the part numbers of the power cables available.
If the plug on the cable does not fit the power outlet, or the cable is to be attached to a terminal block, cut the cable at the plug end and re-wire it. This work should be performed by a qualified electrician all local electrical codes being strictly observed.
The color coding used in the cable will depend on the cable supplied. If a new plug is to be connected, it must meet local safety requirements and include the following features:
Installation Requirements

Installation Requirements

The Agilent B2200 must be operated within the following environmental conditions:
The Agilent B2200 should be stored or shipped in environments within the following limits:
B2200A/B2210A: < 80 % R.H. (at 35 °C), < 60 % R.H. (at 65 °C), non-condensing
Protect the Agilent B2200 from temperature extremes to prevent condensation from forming inside the Agilent B2200. If condensation occurs, the Agilent B2200 may damage or may not satisfy the specifications. If the Agilent B2200 was placed at the high temperature and high humidity environment long time, the Agilent B2200 may degrade its performance.
The Agilent B2200 must be packed in the certain packing materials for protection from damage when it is shipped. Cover the Agilent B2200 by using the moisture-proof and dehumidifying packing material (furnished), and pack it by using packing materials such as carton box and cushioning materials.
Installation Requirements

The Agilent B2200 contains the condensation sensitive electronic parts. The condensation will have a negative impact on the Agilent B2200 to operate normally.
Do not open the packing materials, and leave the Agilent B2200 to acclimate it to the installation environment (temperature and humidity). If it is opened without enough acclimation, the Agilent B2200 may damage.
When you open the box that contains the Agilent B2200, check the following:
If anything is wrong, notify your local Agilent Technologies sales office.
Installation Inspection

Installation Installing the B2200
This section describes how to install the Agilent B2200.
Set the GPIB address. See “To Set the GPIB Address” on page 2-8 and “To Connect the GPIB Cable” on page 2-8.
Turn the Agilent B2200 off, remove the power cable, and install the switch module. See “To Install the Switch Module” on page 2-9 and “To Install the Blank Panel” on page 2-9.
If you require that the modules be guaranteed to their specifications (for ISO compliance, etc.), then you must perform a calibration on the instrument (modules and mainframe together). For calibration, contact your nearest Agilent Technologies service center.

Installation Installing the B2200
Every device on the GPIB bus must have a unique address. If you need to change the GPIB address, turn the Agilent B2200 on and perform the following procedure. The new GPIB address is recognized only at power on. The Agilent B2200 leaves the factory with the GPIB address set to 22.
GPIB Address = XX
To connect the instrument with a computer or peripheral device via GPIB (IEEE Std. 488), connect an GPIB cable between the GPIB connector on the instrument rear panel and the GPIB connector on the peripheral device.
Including the controller, a total of 15 GPIB interfaces can be connected on the same GPIB bus. The following are rules for connecting GPIB interfaces:
Max. total cable length = number of interfaces × 2 m.
Max. total cable length = 20 m.

Installation Installing the B2200
Use clean handling and anti-static procedures when removing, configuring, and installing the switch modules. The modules contain components that can be damaged by static electricity.
The following procedure explains the module installation and removal.
If the blank panel is not installed to cover an unused slot, install the blank panel as follows:

The self-test and diagnostics checks the operation of the mainframe and the modules. However they cannot confirm if the Agilent B2200 satisfies its specifications.
For verifying the specifications, contact your nearest Agilent Technologies Service Center. Trained service personnel will perform calibration (performance verification).
It is recommended to perform calibration once a year at least.
The following procedure performs the self-test and diagnostics.
The controller test will be performed. If the Agilent B2200 fails the test, contact your nearest Agilent Technologies Service Center.
KEY Front panel interface test
LED LED matrix test
PEN Light pen test
To display the previous test result, move the cursor to RESULT, then press Enter key.
For the details of each test, see “Selftest Menu” on page 3-34.

This section describes how to connect the Agilent B2200 outputs to prober, connector plate, test fixture, and so on (DUT interface).
If you do not use the connector plate for the connection between the output and the DUT interface, see “To Mount Connectors” on page 2-18.
The Agilent B2200 output connectors are the triaxial BNC connector. The input signals appears at the output as shown in Figure 2-1 when input-output switching path is made. Figure 2-1 shows signals appear on the output connectors with non-Kelvin connection. If you make Kelvin connection, the signal of the even output connectors is Sense, not Force.
Figure 2-1 Output Connector and Output Signal
GND
GND
(a) When SMU is connected (b) When AUX is connected
Installation Output Connections · Signal (force) · Signal (force)

Connector plates (Table 2-2) are used for the connection between the Agilent B2200 outputs and the DUT interface (prober and so on). To connect to the connector plate, use the cable shown in Table 2-3.
Table 2-2 Connector Plate
Agilent Model No. Description
16495F Half size connector plate
16495F-001 has 12 triaxial through connectors (female to female), an Intlk connector, and a GNDU connector (triaxial through, female to female). The back of the Intlk connector is designed for soldering.
16495F-002 has 12 triaxial connectors, an Intlk connector, and a GNDU connector. The back of each connector is designed for soldering.
16495G Full size connector plate
16495G-001 has 24 triaxial through connectors (female to female), an Intlk connector, and a GNDU connector (triaxial through, female to female). The back of the Intlk connector is designed for soldering.
16495G-002 has 24 triaxial connectors, an Intlk connector, and a GNDU connector. The back of each connector is designed for soldering.
This plate is used to cover openings when you made too big openings for mounting the connector plate. You will use this plate to cover openings if you mount the half size connector plate in openings made for the full size connector plate.
Table 2-3 Output Cable
Agilent Model No. Description
16494A Triaxial cable (for non-Kelvin connection)
16494B Kelvin Triaxial cable (for Kelvin connection)
To install the connector plate, refer to Agilent 16495 Installation Guide.
Installation Output Connections · 16495E Blank plate

For Kelvin connection, use Kelvin triaxial cable listed in Table 2-3. To make a Kelvin output port (1, 3, 5, 7, 9 and 11), couple two E5252A output ports as follows:
Kelvin Output Port Output Port Number
1 1 (Force) and 2 (Sense)
3 3 (Force) and 4 (Sense)
5 5 (Force) and 6 (Sense)
7 7 (Force) and 8 (Sense)
9 9 (Force) and 10 (Sense)
11 11 (Force) and 12 (Sense)
This section describes for the connections between the DUT interface and the connectors connected to the Agilent B2200 output cables. See Table 2-4.
For the extended measurement paths over the connector plate, use low-noise coaxial cable (Agilent part number 8121-1191). This cable can maximize the guard effects and minimize the impression of the external noise.
Figure 2-2 shows the cutting example of this cable. Key point is the isolation between the conductive layer and the center conductor. So, cut and trim the end of the cable as shown in this figure by using a cutter and so on.
Figure 2-2 Coaxial Cable Cutting Example
Cover here using sleeve
Center conductor (for Force/Sense signal)
approx. 10 to 15 mm
Outer conductor (for Guard signal) Insulator (clear)

Table 2-4 Connection to the DUT Interface

This connection is available only for the Kelvin connectors.
Following connection is for the Kelvin connectors. For the triaxial connectors, ignore SENSE terminal, and make connection only for FORCE terminal. Measurement data will include residual resistance from the connection cable.
This connection can cancel effects of cable resistance by connect sense line and force line as close as possible to DUT terminal.
For highly accurate current forcing and measurements while minimizing leakage, surround all force and sense lines from SMU by a guard as far as possible, and make cables stable by taping.
For reducing capacitance measurement error, tape the cables to any grounded materials such as shielding box. If you use probe card, grounding the probe card also reduces capacitance measurement error.
Installation Output Connections

The interlock circuit is to prevent electric shock when touching measurement terminals. You must install an interlock circuit on shielding box to prevent dangerous voltages when door of the shielding box is open.
Figure 2-3 shows the pin assignments of the interlock connector.
Figure 2-3 Interlock Connector Pin Assignments
Install the interlock circuit as follows.
For example, Agilent 4155/4156 is connected to the interlock circuit, it cannot force more than ±40 V when the door is open. When door is closed, it can force more than ±40 V.
Installation Output Connections


Figure 2-4 Dimensions of Interlock Switch (Agilent part number 3101-0302)
Figure 2-5 Dimensions of Interlock Switch (Agilent part number 3101-3241)

To install LED circuit on your shielding box, do following:

The 4155/4156 semiconductor parameter analyzer's Intlk connector provides the interlock signal and a LED drive signal. If a LED is connected between pin 4 and pin 5 (or 6) of the interlock connector, the LED lights to indicate high voltage output when more than ±40 V is forced from an SMU in the 4155/4156.
Figure 2-6 Dimensions of LED (Agilent part number 1450-0641)
Before beginning the measurement, connect the interlock circuit to the instrument’s interlock connector as follows.
Connect the 16493J interlock cable between the interlock circuit and the instrument’s interlock connector. You can connect it directly without using any adapter.
Installation Output Connections · Cathode(-) · Anode(+)

If you use the 4155/4156, you can easily check the interlock circuit as follows:
To stop the interlock test, select STOP secondary softkey.
This section provides the information needed to mount connectors for the Agilent B2200 output cable connections on the shielding panel. See this section if you do not use the connector plate.
For kelvin connections, use the Kelvin cables. The Kelvin cable requires the kelvin triaxial connector which has two connector holes and three screw holes.
Installation Output Connections

Table 2-5 Recommended Parts
Description Agilent Part No.
Interlock Connector (6 pin, female) 1252-1419C
Switch 3101-0302 or 3101-3241
LED (VF @ 2.1 V @ IF = 10 mA) 1450-0641
Triaxial Connector (female) 1250-2457
Low Noise Coaxial Cable 8121-1191
Table 2-6 Dimensions of Connector Holes
Triaxial Connector (in mm) Interlock Connector (in mm)
∅11.3
Kelvin Triaxial Connector (in mm)
14 11 11 14 2.8
2 − ∅11.3 ∅3.2 2 − M3 x 0.5
Installation Output Connections · Wire 8150-5680

This section explains how to connect instruments to the Agilent B2200.
The Agilent B2200 input connector has eight SMU input connectors and six AUX input connectors. Table 2-7 shows cables used to connect the instrument for typical applications using the Agilent B2200.
You can make the Kelvin connections when using the E5260/E5270/4142/4156 SMU or 41501 HPSMU. Then use cables as follows:
To connect the Kelvin triaxial cable to the Agilent B2200 input, use pairs (couple ports) of the SMU input connectors (1&2, 3&4, 5&6, or 7&8). The input signal for each Kelvin input is as follows:
Ground Unit (GNDU) can sink over 1 A. For the measurement over 1 A, connect the GNDU to the connector plate directly by using the Agilent 16493H GNDU cable.
If you never perform the measurement over 1 A, you can connect the GNDU to the Agilent B2200 inputs. Then use the Agilent 16493N GNDU cable. This cable requires the Kelvin connection.
Installation Input Connections

Table 2-7 B2200 Input Connections
Cable Required Agilent Model No.
4155 MPSMU SMU1 to 8 16494A triaxial cable
VSU AUX1 to 6 BNC cable
VMU AUX1 to 6 BNC cable
Intlk − 16493J interlock cable (for direct connection to connector plate)
4156 HRSMU SMU1 to 8 16493K Kelvin triaxial cable or 16494A triaxial cable
VSU AUX1 to 6 BNC cable
VMU AUX1 to 6 BNC cable
Intlk − 16493J interlock cable (for direct connection to connector plate)
41501 HPSMU SMU1 to 8 16493K Kelvin triaxial cable or 16494A triaxial cable
MPSMU SMU1 to 8 16494A triaxial cable
PGU AUX1 to 6 BNC cable
GNDU SMU1 to 8 16493N GNDU Kelvin cable
− 16493H GNDU cable (for direct connection to connector plate)
4142B SMU SMU1 to 8 16494B Kelvin triaxial cable or 16494A triaxial cable
VSU AUX1 to 6 BNC cable
VMU AUX1 to 6 BNC cable
GNDU SMU1 to 8 16493HN GNDU Kelvin cable
− 16493H GNDU cable (for direct connection to connector plate)
Intlk − 16435A interlock cable adapter and 16493J interlock cable (for direct connection to Connector Plate)

Cable Required Agilent Model No.
SMU SMU1 to 8 16493K Kelvin triaxial cable or 16494A triaxial cable
GNDU SMU1 to 8 16493N GNDU Kelvin cable
− 16493H GNDU cable (for direct connection to connector plate)
Interlock − 16493J interlock cable (for direct connection to connector plate)
C Meter High, Low CMH/CML 16494F CMU cable for 4 terminal pairs
Pulse Input Pulse Generator
Output AUX1 to 6 BNC cable
Bias Input Power Supply
Output AUX1 to 6 BNC cable
In the table above,

Installation Measurement Cable Length
This section describes how to calculate the total guard capacitance when using an SMU (source monitor unit). When using an SMU, the length of measurement cables is limited by the guard capacitance of the cables. The guard capacitance means the capacitance between the signal line (Force or Sense) and the Guard line.
Table 2-8 lists the guard capacitance for each element of the Agilent B2200’s measurement environment.
When using the 4155/4156/4142B/E5260/E5270, the maximum limit of the guard capacitance is approximately 900 pF. So, you add the following and total must be less than 900 pF:
The following is an example to calculate the guard capacitance of a measurement environment:
Cable from Connector Plate to DUT: 8121-1191, 1 m (130 pF)
In this environment, the total guard capacitance (Cg-total) is as follows.
Cg-total = 145+(8×3)+75+130+130+10 pF = 514 pF < 900 pF
Even if you use other instruments, you can use the table for reference about the measurement cable length and capacitance.

Installation Measurement Cable Length
Table 2-8 Guard Capacitances of B2200 Measurement Environment (Typical)
Connection Agilent Model/Part No. Cable Length
from SMU to B2200 inputs
16494A-003 (for non-Kelvin) 80 cm 75 pF
16494A-001 (for non-Kelvin) 1.5 m 130 pF
16494B-003 (for Kelvin, 4142B) 80 cm 90 pF
16493K-001 (for Kelvin) 1.5 m 150 pF
B2210A (only 1 card is installed) − 145 pF
B2210A (more 1 card) − 8 pF/extra card
B2211A (only 1 card is installed) − 145 pF
B2211A (more 1 card) − 8 pF/extra card
from B2200 outputs to connector plate or probe card interface
16494A-001 (for non-Kelvin) 1.5 m 130 pF
16494A-002 (for non-Kelvin) 3 m 240 pF
16494A-003 (for non-Kelvin) 4 m 340 pF
16494B/C-001 (for Kelvin) 1.5 m 140 pF
16494B/C-002 (for Kelvin) 3 m 260 pF
16494C-003 (for Kelvin) 4 m 340 pF
from connector plate to DUT
Agilent Part No. 8121-1191 X m 130 pF per m
probe card interface B2220A − 70 pF
Guard Capacitance · mainframe and modules

Maintenance should be performed periodically to keep the B2200 in good condition.
Calibration must be performed periodically so that the instruments satisfy the specifications, and keep a good condition. It is recommended to perform a calibration once a year at least. For calibration, contact your nearest Agilent Technologies Service Center. Trained service personnel will perform calibration (performance verification).
Before performing cleaning, turn off the instrument, and disconnect power cable from the rear panel. Use a dry cloth to clean the external case parts.
To prevent electrical shock, do not perform cleaning when the instrument is turned on, and do not use a wet cloth.
Installation Maintenance

Installation Maintenance


This chapter explains the front panel operation and the switch control functions of the Agilent B2200, also provides the reference information of the front panel keys and display.
Front Panel Operation

This section describes operations of the Agilent B2200.
Use the right/left arrow key to move the cursor to the desired function name. Use the up/down arrow key to select the value for the setup item.
The Enter key makes the setup effective, and returns the display to the previous menu or display.
The Exit key returns the display to the previous menu or display without making the setup effective.
Front Panel Operation · Operation

Front Panel Operation Operation
Turn the Agilent B2200 off, and connect the light pen to the Light Pen connector at the right down corner of the front panel. After that, perform the following procedure to enable the light pen.
You can control the Agilent B2200 functions and the switch connections by using the front panel keys. And the Agilent B2200 allows you to use the Agilent 16443A Light Pen to control the switch connections.
Pointing a LED on the LED matrix by using the light pen controls the setting of the switch specified by the LED.
Changing the channel configuration mode resets the Agilent B2200.

If you use the light pen, you do not need the following instruction to control the switch condition.
Open/Close key. The blink-LED will appear.
Blinking of LED will be stopped by pressing the front panel key other than the
Open/Close, Fast, or arrow keys.
For the Kelvin outputs, the input-output paths must be controlled as shown below.
Control the switches to connect the input couple port to the Kelvin output.
Control the switches to connect the input port to two ports of the Kelvin output.
If you do not comply with this note, leakage current between the Kelvin paths will cause measurement error.
Front Panel Operation · Operation

Front Panel Operation Operation
When the bias mode is ON, the bias-enabled output ports will be connected to the input bias port.
When the ground mode is ON, the ground-enabled input/output ports will be connected to the input ground port.
The Agilent B2200 provides the internal memory used to save the setup condition. Eight setup conditions can be saved into the memory.
Wait until the save operation is completed. Then, the LCD displays the status and setup conditions.

Bias mode cannot be set to ON when the ground mode is ON.
a. Press the Shift key and the Port Function key.
b. Move the cursor to SET, then press the Enter key.
c. Move the cursor to DISABLE, then press the Enter key.
The LCD displays one of the bias-enabled output port numbers which can be changed to the bias-disabled. Or if there is no bias-enabled output port, the LCD displays Channel: No Channel. Press the arrow key to select the output port to be changed to the bias-disabled, then press the Enter key.
d. Move the cursor to ENABLE, then press the Enter key.
The LCD displays one of the bias-disabled output port numbers which can be changed to the bias-enabled. Or if there is no bias-disabled output ports, the LCD displays Channel: No Channel. Press the arrow key to select the output port to be changed to the bias-enabled, then press the Enter key.
e. Repeat c and d until the bias-enabled port setup is completed. After that, press the Exit key more than once to display the status and setup conditions.
a. When the bias mode is ON, press the Port Function key. The blink-cursor appears on the first line of the LCD.
b. Press the right/left arrow key to move the cursor on the input port number
(01 to 14) that will be set to the input bias port.
c. Press the up/down key to display B, then press the Enter key.
Front Panel Operation · Operation

Front Panel Operation Operation
Ground mode cannot be set to ON when the bias mode is ON.
a. Press the Shift key and the Open All key.
b. Move the cursor to SET, then press the Enter key.
c. Move the cursor to ENABLE, then press the Enter key.
The LCD displays one of the ground-disabled output port numbers which can be changed to the ground-enabled. Or if there is no ground-disabled output ports, the LCD displays Channel: No Channel. Press the arrow key to select the output port to be changed to the ground-enabled, then press the Enter key.
d. Move the cursor to DISABLE, then press the Enter key.
The LCD displays one of the ground-enabled output port numbers which can be changed to the ground-disabled. Or if there is no ground-enabled output port, the LCD displays Channel: No Channel. Press the arrow key to select the port to be changed to the ground-disabled, then press the Enter key.
e. Repeat c and d until the ground-enabled port setup is completed. After that, press the Exit key more than once to display the status and setup conditions.
a. When the ground mode is ON, press the Port Function key. The blink-cursor appears on the first line of the LCD.
b. Press the right/left arrow key to move the cursor on the input port number
(01 to 14) that will be set to the input ground port.
c. Press the up/down key to display G.
a. Press the right/left arrow key to move the cursor on the input port number (01 to 08) that will be set to the ground-enabled input port.
b. Press the up/down key to display -.
Open the ground-enabled input ports to prevent the instrument from damage.

a. Press the Menu key. The setup menu will be displayed on the LCD.
b. Move the cursor to SCAN, then press the Enter key. The LCD displays NO.
c. Press the arrow key to display YES, then press the Enter key.
d. Press the Exit key.
a. When the couple mode is ON, press the Port Function key. The blink-cursor appears on the first line of the LCD.
b. Press the right/left arrow key to move the cursor on the input port number
(01 to 14) that will be set to the input couple port.
c. Press the up/down key to display C.
d. Repeat b and c for the all couple ports, then press the Enter key.
Front Panel Operation · Operation

Front Panel Operation Operation

This instruction enables or disables the data display in the GPIB remote condition. See “RMT_DSPL” on page 3-33.
Press the Local key. If the front panel keys are locked, send the :SYST:KLC command from an external computer, then press the Local key.
Front Panel Operation · Operation

Front Panel Operation Switch Control Functions
This section introduces the switch control functions of the Agilent B2200.

Front Panel Operation Switch Control Functions
The Agilent B2200 provides the two channel configuration modes, Normal and Auto. The configuration mode defines the way to control multiple switch modules installed in the mainframe.
Normal Each module is considered independently. So each module is always a 12 output switching matrix.
The switch control functions need to be defined for each module.
Auto Default mode at Power On or *RST command.
To use this mode, the same type of module must be installed in the mainframe from slot number 1 continuously (slots 1 to 2; or slots 1 to 3; or slots 1 to 4).
The installed modules are treated as one switching matrix. For example, if the modules are installed in slots 1 to 3, the Agilent B2200 works as a 36 output switching matrix.
The switch control functions are available for the switching matrix.

Front Panel Operation Switch Control Functions
Connection rule is used to specify if an input or output port can have multiple connections. For each module installed in the mainframe (for auto configuration mode, installed modules are treated as one module), you can specify one of the following connection rules:
Single: Each input port can be connected to only one output port, and each output port can be connected to only one input port. So, existing connection to a port will be disconnected when a new connection is made.
If normal configuration mode is set to the Agilent B2200 with multiple modules, an input/output path is effective for each module. For example, if normal configuration mode is set to the Agilent B2200 with two modules, and single rule is set to both modules, the following connection is possible.
1 2 12 3 3
Free: Each input port can be connected to multiple output ports, and each output port can be connected to multiple input ports.

Front Panel Operation Switch Control Functions
Connection sequence specifies the open/close sequence of the relays when changing from an existing connection to a new connection.
You can select one of the three connection sequences shown in Table 3-1 for each module installed in the Agilent B2200.
When an existing connection is changed to a new connection, the relays are opened/closed as follows:
Table 3-1 Connection Sequence
Break Before Make 1. Disconnect previous route. 2. Wait for relays to open. 3. Connect new route.
Make Before Break 1. Connect new route. 2. Wait for relays to close. 3. Disconnect previous route.
No Sequence 1. Disconnect previous route. 2. Connect new route.
Connection · Sequence Relay Operation

Front Panel Operation Switch Control Functions
Bias mode is useful for connecting the same input to multiple channels (output ports) simultaneously. And the connections will be kept until the other input port connection will be changed.
When the bias mode is ON, the input bias port is connected to all bias enabled output ports that are not connected to any other input ports. You cannot directly control which output ports are connected to the input bias port:
When the bias mode is OFF, the input bias port is the same as the other input ports, so relays can be controlled directly to connect to output ports.
Figure 3-1 shows an example. Input 2 is the input bias port. When the bias mode is turned ON, all bias enabled output ports (that are not connected to other input ports) are connected to the input bias port. So, the output ports that are connected to input port 1 and 3 are not connected to the input bias port.
Figure 3-1 Example of Bias Mode

Open Connects to bias port Connects to input 01

Front Panel Operation Switch Control Functions
Connection sequence (to connect input bias port to output ports) is always Break-Before-Make.
Bias mode cannot be set to ON when the ground mode is ON.
If the bias input port and a couple port have been assigned to the same input port, the bias mode and the couple mode cannot be used in parallel.

Front Panel Operation Switch Control Functions
Ground mode operation is similar to the bias mode operation. The ground mode is useful for connecting the same input to multiple channels (output ports) and the unused input ports simultaneously. And the connections will be kept until the other input port connection will be changed. The ground mode is especially useful for settling the potential of the unused input/output paths.
If you assign the input port 12 (AUX Input 12) as the ground port, open the input port 12. This ground port will be internally connected to the ground when the ground mode is set to ON.
When the ground mode is ON, the input ground port is connected to all ground enabled input ports (unused ports) and the ground enabled output ports that are not connected to any other input ports. You cannot directly control which output ports are connected to the input ground port:
When the ground mode is OFF, the input ground port is the same as the other input ports, so relays can be controlled directly to connect to output ports.
Connection sequence (to connect input ground port to output ports) is always Break-Before-Make.
Ground mode cannot be set to ON when the bias mode is ON.
The input ground port and a ground enabled input port cannot be assigned to the same input port.
If the ground input port and a couple port have been assigned to the same input port, the ground mode and the couple mode cannot be used in parallel.
If a ground enabled input port and a couple port have been assigned to the same input port, the ground mode and the couple mode cannot be used in parallel.

Front Panel Operation Switch Control Functions
Open the ground enabled input ports (unused input ports). If any equipment is connected to the ground enabled input ports, turning the ground mode ON may cause damage in the equipment.

Front Panel Operation Switch Control Functions
Couple mode is useful for making the Kelvin connections. When the couple mode is ON, the input couple ports will be connected to the output couple ports. For the available couple ports, see Table 3-2.
Table 3-2 Input Couple Port Numbers and Output Couple Port Numbers
Available input couple port No. Input ports Available output couple port No.
1 SMU 1, 2 1 and 2, 3 and 4, . . . , 11 and 12 (for normal configuration mode)
1 and 2, 3 and 4, . . . , 47 and 48 (for auto configuration mode)
7 SMU 7, 8
9 AUX 9, 10
11 AUX 11, 12
13 AUX 13, 14 (CMH, CML)
For example, if you specify the SMU input 1 for the couple port and set the couple mode ON, the input 1 will be connected to an odd number of the output port and the input 2 will connected to the next number of the output port. And if you specify to connect the input 1 to the output 10, the input 1 will be connected to the output 9 and the input 2 will be connected to the output 10.
If the ground input port and a couple port have been assigned to the same input port, the ground mode and the couple mode cannot be used in parallel.
If a ground enabled input port and a couple port have been assigned to the same input port, the ground mode and the couple mode cannot be used in parallel.

Front Panel Operation Switch Control Functions
The Agilent B2200 provides the function to detect the input ports that connect the Kelvin triaxial cable and set the ports as the couple port automatically. See “SCAN” on page 3-32

Front Panel Operation Display Functions
The Agilent B2200 provides LED matrix, LCD, and 18 front panel keys for front panel operation. This section explains the display of the LED matrix and the LCD.
Agilent B2200 has four blocks of 14 × 12 LED matrix. They display the switch condition of the switch modules installed in the slot 1 to slot 4. See Table 3-3.
Also LEDs labeled Card 1 to Card 4 are located above the LED matrices. They indicate the status of the module installed in the slot 1 to 4, respectively. See Table 3-4.
Table 3-3 LED Matrix
Green The switch has been closed to connect the output channel to the input port.
Red The switch has been closed to connect the output channel to the input bias port or the input ground port.
Orange The LED also blinks. The switch condition can be changed by the
Open/Close key.
- The LED lights out. The switch has been opened.
Table 3-4 Card 1/2/3/4 LED
Red The switch module has failed the selftest or the diagnostics.
Green The switch module has operated normally.
- The LED lights out. No switch module has been installed in this slot.
LED color Explanation · LED color Explanation

The LCD displays the setup information and the status information as shown in Table 3-5. Display example is shown in Figure 3-2.
Figure 3-2 LCD Display Example
01 02 03 04 05 06 07 08 09 10 11 12 13 14 Shift
Sequence Rule Gnd Lock Error Gnd Bias
The first line shows the input port status, couple port, bias port, ground port, or ground enabled port.
The second line shows the instrument status and the mode status.
The example of Figure 3-2 notifies you the following information.
Front Panel Operation · Display Functions

Front Panel Operation Display Functions
Table 3-5 LCD Display Items
01 to 14 Port function assigned to the input ports 1 to 14. B (bias port), C (couple port), G (ground port), - (ground enabled port), or blank (no function).
Shift Shift key status indicator. The triangle mark appears when the sub key is active.
Remote Remote status indicator. The triangle mark appears when the Agilent B2200 is in the GPIB remote condition.
Lock Key status indicator. The triangle mark appears when the front panel keys are locked by an external computer.
Error Error status indicator. The triangle mark appears when any error has occurred.
Card# Slot number of the switch module to be monitored. Displays A when the channel configuration mode is Auto. In the auto mode, the modules installed in the mainframe from slot number 1 continuously will be controlled as one module. Or displays 1 to 4 when the channel configuration mode is Normal. The module can be selected by the Card Selection key.
Bias Bias mode status indicator. The triangle mark appears when the mode is ON.
Couple Couple mode status indicator. The triangle mark appears when the mode is ON.
Gnd Ground mode status indicator. The triangle mark appears when the mode is ON.
Rule Connection rule, FREE (plural route mode) or SINGLE (single route mode).
Sequence Connection sequence, BBM (Break Before Make), MBBR (Make Before Break), or NO_SEQ (No Sequence).
Label Description


The Agilent B2200 provides LED matrix, LCD, and 18 front panel keys for front panel operation. The front panel keys are used to change the instrument settings, the switch conditions, and so on.

Local Sets the Agilent B2200 to the local condition.
Used to reset the Agilent B2200.
Displays the following message. Press the arrow key to set YES, and then press the
Enter key to reset the instrument.
Menu Displays the setup menu. See “Setup Menu” on page 3-30.
Displays the selftest menu. See “Selftest Menu” on page 3-34.
Reset (Shift+Local) · Selftest (Shift+Menu)

Front Panel Operation Front Panel Keys
The Function key group is used to change the input ports and modes.

Port Function Enters into the port function selection mode. In this mode, the arrow keys, Fast key,
Exit key, and Enter key are available. Other keys will work to exit this mode.
In this mode, move cursor right or left to specify the input port, and press the up or down arrow key to select the port function, B (bias port), C (couple port), G (ground port), or - (ground enable port). The port function can be selected when the corresponding port operation mode (bias mode, couple mode, or ground mode) is set to ON. The bias mode and the ground mode cannot be set to ON simultaneously.
Bias Ch (Shift+ Port Function)
Displays the following menu.
The function displays the following message for example. And then press Enter to bias-enable or bias-disable the channel.
If there is no enabled or disabled channel, the following message will appear.

Bias Mode Sets the bias mode ON or OFF.
When the bias mode is set to ON, the LCD shows which input port is the bias port, and the input bias port will be connected to the bias-enabled channels that are not connected to other input port. And LEDs will be turned red to show the bias port connections.
Ch Mode (Shift+ Bias Mode)
Sets the channel configuration mode AUTO or NORMAL.
Changing the mode will reset the Agilent B2200 settings.
Couple Mode Sets the couple mode ON or OFF.
When the couple mode is set to ON, the LCD shows which input ports are the couple port.
Gnd Mode (Shift+ Couple Mode)
Sets the ground mode ON or OFF.
When the ground mode is set to ON, the LCD shows which input port is the ground port, and the input ground port will be connected to the ground-enabled channels that are not connected to other input port. And LEDs will be turned red to show the ground port connections. The ground enabled input ports are also connected to the input ground port. Nothing must be connected to the ground enabled input ports.
Rule Displays the following message. Selects the connection rule FREE (plural route connections) or SINGLE (single route connection).
Displays the following message. Selects the connection sequence Break_Before_Make, Make_Before_Break, or No_Sequence.
Load Memory Displays the following message. Used to select the Agilent B2200 setup data stored in the internal memory, and loads the data.
Save Memory (Shift+ Load Memory)
Displays the following message. Used to select the internal memory to store the Agilent B2200 setup data, and stores the data. Up to eight setup data can be saved.
Open All Displays the following message. Press the arrow key to set YES, and then press the
Enter key to open all relays.

Front Panel Operation Front Panel Keys
Gnd Ch (Shift+Open All)
Displays the following menu.
The function displays the following message for example. And then press Enter to ground-enable or ground-disable the channel.
If there is no enabled or disabled channel, the following message will appear.

The Edit key group is used to select the switch module to be monitored, move the cursor in the display, make the setup value effective, return to the previous menu or display, and so on.

Card Selection Selects the switch module to be monitored on the LCD. The Card# status indicator will change the slot number (card number) when the Card Selection key is pressed.
This key is not effective when the Card# indicator shows A (auto configuration mode).
Shift Enables the sub key function. The sub key label has been printed on the front panel above the key. The label color is blue.
Exit Returns the display to the previous menu or display without making the setup effective.
Enter Makes the setup effective, and returns the display to the previous menu or display.
Fast Used with the arrow key. Pressing the Fast key and an arrow key speeds up changing the setting value.
arrow keys Moves the cursor on the display, displays another selection for a setup message, or changes the value of the digit specified by the cursor.
For the LED matrix, moves the blink-LED.
Open/Close When no blink-LED is on the LED matrix, enters into the LED matrix based switch control mode. The blink-LED will appear. In this mode, the arrow keys and Fast key are available. Other keys will work to exit this mode.
When the blink-LED is on the LED matrix, changes the condition, open or close, of the switch specified by the blink-LED.
Front Panel Operation · Front Panel Keys

Front Panel Operation Setup Menus
This section explains the following setup menus displayed by pressing the Menu key or Selftest key.
Press the Menu key. The setup menu will be displayed on the LCD as shown in the following example. The setup menu is used to change the GPIB address, detect the couple port automatically, and so on.
[Hardware Configuration] CONFIG SCAN ERROR RMT_DSPL BEEP >
[Hardware Configuration]
Press the arrow key to move the cursor to the desired function name. Then, press the
Enter key. The setup message or the sub menu will be displayed. To return to the previous menu, press the Exit key.

Function Tree The setup menu provides the following functions.
Front Panel Operation · Setup Menus

Front Panel Operation Setup Menus
CONFIG Displays the sub menu that provides the following functions.
Displays the following message.
GPIB Address = Address
Press the arrow key to set the desired GPIB address. Then, press the Enter key to make the setup effective, or press the Exit key to cancel changing the setup.
Displays the following message.
where X.XX.XX shows the firmware revision.
To return to the previous menu, press the Exit key.
Displays the following message.
where # is 1, 2, 3, or 4, model is the model number of the module, X is the revision of the module.
To display another message, press the arrow key.
To return to the previous menu, press the Exit key.
SCAN Displays the following message.
Press the arrow key to set YES. Then, press the Enter key to detect and set the input couple ports automatically, or press the Exit key to cancel this operation. The input couple ports will be the input ports the Kelvin cable is connected.

ERROR Displays the sub menu that provides the following functions.
Displays the error code and error message, or No Error.
To return to the previous menu, press the Exit key.
Displays the following message. Press the arrow key to set YES. Then, press the
Enter key to clear the error buffer, or press the Exit key to cancel this operation.
RMT_DSPL Displays the sub menu that provides the following functions.
Displays the following message. Press the arrow key to set ON (displays all status information in the GPIB remote mode) or OFF (displays Remote, Lock, and Error only). Then, press the Enter key to make the setup effective, or press the Exit key to cancel changing the setup.
Displays the following message. Press the arrow key to set ON (enables LED matrix in the GPIB remote mode) or OFF (disables it). Then, press the Enter key to make the setup effective, or press the Exit key to cancel changing the setup.
BEEP Displays the following message. Press the arrow key to set the beeper ON or OFF. Then, press the Enter key to make the setup effective, or press the Exit key to cancel changing the setup.
PEN Displays the following message. Press the arrow key to enable or disable the light pen. Then, press the Enter key to make the setup effective, or press the Exit key to cancel changing the setup.

Front Panel Operation Setup Menus
Press the Shift key and the Menu key. The selftest menu is displayed. The menu is used to execute the selftest and diagnostics.
[Execute Diagnostics] SELF_TEST RELAY_TEST KEY BEEPER >
[Execute Diagnostics]
Press the arrow key to move the cursor to the desired function name. Then, press the
Enter key. The setup message or the sub menu will be displayed. To return to the previous menu, press the Exit key.
Function Tree This menu provides the following functions.

SELF_TEST Displays the sub menu that provides the following functions.
Starts the controller test. Wait until PASS or FAIL is displayed. Then press the
Exit key to return to the previous menu.
Displays the test result. Press the Exit key to return to the previous menu.
RELAY_TEST Displays the sub menu that provides the following functions.
Before starting the relay test, disconnect cables from the input connectors. And open the outputs (end of cable is OK). For the Kelvin cable, disconnect it from the output connectors.
Displays the following message.
Press the up or down arrow key to select the module to be tested. Then press the
Enter key to start the relay test, or press the Exit key to cancel this operation.
Number of modules is 2 or more, the relay test will be automatically completed. However, if the only one module is installed, the following message will appear.
Connect IV1 and 2. Press [Enter].
Then connect a cable between the input 1 and 2, and press Enter key.
After that, change the cable connection as shown in the displayed message and repeat this for the other input connections (input 3 and 4, 5 and 6, 7 and 8, 9 and 10, 11 and 12, 13 and 14). Do not forget to open the connectors other than the specified pair.
Displays the following message (example). Press the up or down arrow key to display the test result for each module. Press the Exit key to return to the previous menu.
Front Panel Operation · Setup Menus · Slot1

Front Panel Operation Setup Menus
KEY Displays the sub menu that provides the following functions.
Displays the following message.
DIAG:KEY Press any key.
Press any front panel key and confirm the LCD display. For example, if you press the Port Function key, the following message must be displayed.
DIAG:KEY [Port Function]
Continue this for all keys. After that, press the Enter key twice for the normal operation. The test result will be PASS.
If you find any wrong response, press the Exit key twice. The test result will be FAIL.
Finally, press the Exit key to return to the previous menu.
Displays the test result. Press the Exit key to return to the previous menu.
BEEPER Displays the sub menu that provides the following functions.
Starts the beeper test, and displays the following message.
DIAG:BEEP Is beeper making 2 sounds?
Press the Enter key if the beeper works correctly. The test result will be recorded as PASS.
If you find any problem, press the Exit key. The test result will be recorded as FAIL.
Finally, press the Exit key to return to the previous menu.
Displays the test result. Press the Exit key to return to the previous menu.

LED Displays the sub menu that provides the following functions.
Starts the LED matrix test, and displays the following message.
DIAG:LED All LED in Orange?
Press the Enter key if the all LEDs are lighted in orange. The test result will be recorded as PASS.
If you find any problem, press the Exit key. The test result will be recorded as FAIL.
Finally, press the Exit key to return to the previous menu.
Displays the test result. Press the Exit key to return to the previous menu.
PEN Displays the sub menu that provides the following functions.
Before starting the light pen test, connect the light pen to the Agilent B2200.
Starts the light pen test, and displays the following message.
DIAG:PEN Point any cross point
Point the light pen to the point you desire and confirm the LCD display. For example, if you point the input 5-to-output 10 cross point of the Agilent B2200 in the auto configuration mode, the following message must be displayed.
Continue this for all cross points. After that, press the Enter key for the normal operation. The test result will be recorded as PASS.
If you find any wrong response, press the Exit key. The test result will be recorded as FAIL.
Finally, press the Exit key to return to the previous menu.
Displays the test result. Press the Exit key to return to the previous menu.
Front Panel Operation · Setup Menus

Front Panel Operation Setup Menus
GPIB Displays the sub menu that provides the following functions.
Displays the following message.
Open GPIB, then press [Enter]
Disconnect the cable from the GPIB connector on the rear panel, and open it.
Press the Enter key to start the GPIB test, and wait until PASS or FAIL is displayed. Then press the Exit key to return to the previous menu.
Displays the test result. Press the Exit key to return to the previous menu.


This chapter describes the automatic control programming of the Agilent B2200.
Explains the commands for the fundamental switch control of the Agilent B2200.
Provides some examples of the programming.
Explains how to use the capacitance compensation routine. The routine is a function of the Agilent B2200 VXIplug&play driver.
For the details about SCPI commands, see Chapter 5.
For the details about VXIplug&play driver, see Chapter 6.
Programming

This section explains the commands used for the fundamental switch control of the Agilent B2200.
The SCPI commands use a hierarchical structure for subsystem commands similar to a file system. For example, in :ROUT:CONN:RULE command, the hierarchy is as follows:
The colon at the beginning of the command means root. The colons between two command keywords means moving down to a lower level.
A semicolon enables two commands to be sent on the same line. For example, :ROUT:CONN:RULE ALL,FREE;SEQ ALL,BBM is the same as the following two commands:
:ROUT:CONN:SEQ ALL,BBM
So, using a semicolon reduces typing and simplifies the program. A command terminator (such as carriage return) resets the path to root.

The following commands are used to set the fundamental switch control functions of the Agilent B2200. The commands should be entered before performing the open/close operation. For the functions, see “Switch Control Functions” on page 3-12.
Table 4-1 Fundamental Commands
Sets the channel configuration mode.
:ROUT:FUNC ACON
Sets the connection rule. :ROUT:CONN:RULE card_no,FREE
:ROUT:CONN:RULE card_no,SROU
Sets the connection sequence. :ROUT:CONN:SEQ card_no,NSEQ
:ROUT:CONN:SEQ card_no,BBM
:ROUT:CONN:SEQ card_no,MBBR
Programming Programming Basics · Functions Commands

The following commands are used to control open/close of the specified switch.
Table 4-2 Switch Control Commands
Closes the relays specified by channel_list. :ROUT:CLOS channel_list
Opens the relays specified by channel_list. :ROUT:OPEN channel_list
Opens the all relays on the card specified by card_no.
:ROUT:OPEN:CARD card_no
The channel_list is the parameter which determines the input-output cross points to open/close. Figure 4-1 shows the basic syntax of the channel_list parameter.
In the channel_list, you can specify one or more channels. Each specified channel consists of five digits.
Card No.: One digit. 0 when the channel configuration mode is AUTO. Or the slot number (1 to 4) that installs the switch module when the mode is NORMAL.
Channel No.: First two digits: specifies input port. 01 to 14.
Last two digits: specifies output port. 01 to 12, 01 to 24, 01 to 36, or 01 to 48 for the AUTO mode. It depends on the number of modules installed in the mainframe from slot number 1 continuously. Or 01 to 12 for the NORMAL mode
See Figure 1-4.
Figure 4-1 Syntax of channel_list Parameter
,
Channel No. (4 digits) Card No. (1digit) @ Channel No. (4 digits) Card No. (1digit) ( :
) Channel No. (4 digits) Card No. (1digit) @ Channel No. (4 digits) Card No. (1digit) ( :
Programming Programming Basics · Functions Commands · )

In the channel_list, you can specify multiple channels by using comma (,) or colon (:) as follows:
As shown in last two examples, you can specify a range across input ports and card slots.
In the Auto Configuration mode, you can omit any zeros at beginning of channel. For example, channel 00101 can be expressed by 101.
For the Kelvin outputs, the input-output paths must be controlled as shown below.
Control the switches to connect the input couple port to the Kelvin output.
Control the switches to connect the input port to two ports of the Kelvin output.
If you do not comply with this note, leakage current between the Kelvin paths will cause measurement error.
Programming Programming Basics

This section provides examples of control programs for the Agilent B2200.
Example programs use the Microsoft Visual Basic .NET and the Agilent T&M Programmers Toolkit.
The program examples do not include instrument control routines, so to make a measurement, you need to add the desired routines.
If you assign the input port 12 to the ground port, the input port 12 will be connected to the ground internally when the ground mode is ON. In this case, leave open the input port 12.
Programming Programming Examples

The following example connects instrument output to DUT as shown in Figure 4-2.
Figure 4-2 Input-Output Connection Example

from SMU1 to Output 1 (channel list 10101)
from SMU2 to Output 2 (channel list 10202)
from SMU3 to Output 3 (channel list 10303)
from SMU4 to Output 4 (channel list 10404)
Programming Programming Examples

Table 4-3 Input-Output Connection Example
Imports Agilent.TMFramework '1 Imports Agilent.TMFramework.InstrumentIO
Dim B2200 As New DirectIO("GPIB0::22::INSTR") '6 Dim channels As String = "(@10101,10202,10303,10404)" B2200.WriteLine("*RST") '8 B2200.WriteLine(":ROUT:FUNC NCON") B2200.WriteLine(":ROUT:CONN:RULE ALL,SROU") B2200.WriteLine(":ROUT:CONN:SEQ ALL,BBM") B2200.WriteLine(":SYST:DISP:STR 'Connecting MOSFET AG002201'") B2200.WriteLine(":ROUT:CLOS " & channels) '13 MsgBox("Click OK to start measurement.", vbOKOnly, "") Console.WriteLine("Measurement in progress. . ." & Chr(10)) 'insert the code for measurement
B2200.WriteLine(":ROUT:OPEN:CARD ALL") '17 B2200.Close() MsgBox("Click OK to stop the program.", vbOKOnly, "") Console.WriteLine("Measurement completed." & Chr(10)) End Sub '21 End Module
1 to 6 The above example is for the B2200 of the GPIB address 22 on the interface GPIB0. “GPIB0” is the VISA name. Confirm your GPIB settings, and set them properly.
7 Defines the channel list in the channels variable (string).
8 to 12 Resets the Agilent B2200, sets the Normal configuration mode, the Single connection rule, and the Break_Before_Make connection sequence. And displays the string on the LCD.
13 Closes switches to connect the input-output paths specified by channels.
14 to 16 Opens a message box to confirm the start of measurement. And clicking OK on the message box displays a message on the console window.
Insert measurement control program code at the line 16.
17 Opens all relays on the all modules installed in the Agilent B2200.
18 to 22 Disables the software connection with the Agilent B2200, and opens a message box to confirm the end of the program. Finally, clicking OK on the message box displays a message on the console window.

The following example uses the bias mode.
Figure 4-3 Bias Mode Example
from bias port to all bias enabled output ports
The following output ports will be disconnected from the bias port, and the following input-output connections will be made.
from SMU1 to Output 1 (channel list 00101)
from SMU2 to Output 2 (channel list 00202)
from SMU3 to Output 3 (channel list 00303)
from SMU4 to Output 4 (channel list 00404)
Programming Programming Examples

Table 4-4 Bias Mode Example
Imports Agilent.TMFramework Imports Agilent.TMFramework.DataAnalysis Imports Agilent.TMFramework.DataVisualization Imports Agilent.TMFramework.InstrumentIO Module Module1
Dim B2200 As New DirectIO("GPIB0::22::INSTR") Dim channels As String = "(@101,202,303,404)" '8 B2200.WriteLine("*RST") B2200.WriteLine(":ROUT:FUNC ACON") '10 B2200.WriteLine(":ROUT:CONN:RULE ALL,SROU") B2200.WriteLine(":ROUT:CONN:SEQ ALL,BBM") B2200.WriteLine(":ROUT:BIAS:PORT ALL,10") B2200.WriteLine(":ROUT:BIAS:CHAN:ENAB:CARD ALL") B2200.WriteLine(":ROUT:BIAS:STAT ALL,ON") MsgBox("Click OK to start stress output.", vbOKOnly, "") '16 'insert the code for stress output
B2200.WriteLine(":ROUT:BIAS:STAT ALL,OFF") B2200.WriteLine(":ROUT:CLOS " & channels) '19 MsgBox("Click OK to start measurement.", vbOKOnly, "") Console.WriteLine("Measurement in progress. . ." & Chr(10)) 'insert the code for measurement
B2200.WriteLine(":ROUT:OPEN:CARD ALL") B2200.Close() MsgBox("Click OK to stop the program.", vbOKOnly, "") Console.WriteLine("Measurement completed." & Chr(10)) End Sub '27 End Module
8 Defines the channel list in the channels variable (string).
9 to 12 Resets the Agilent B2200. And sets the Auto configuration mode, the Single connection rule, and the Break_Before_Make connection sequence.
13 to 15 Sets the bias port and the bias enabled output ports. And turns the bias mode ON.
16 to 17 Opens a message box to confirm the start of measurement. And clicking OK on the message box displays a message on the console window.
Insert bias source control program code at the line 17.
18 Turns the bias mode OFF.
19 to 28 Same as the code shown in the lines 13 to 22 of Table 4-3.
Programming Programming Examples · Sub Main() · Line Description

The following example uses the ground mode.
Figure 4-4 Ground Mode Example

from ground port to all ground enabled input/output ports
The following output ports will be disconnected from the ground port, and the following input-output connections will be made.
from SMU1 to Output 1 (channel list 00101)
from SMU2 to Output 2 (channel list 00202)
from SMU3 to Output 3 (channel list 00303)
from SMU4 to Output 4 (channel list 00404)
Programming Programming Examples

Table 4-5 Ground Mode Example
Imports Agilent.TMFramework Imports Agilent.TMFramework.DataAnalysis Imports Agilent.TMFramework.DataVisualization Imports Agilent.TMFramework.InstrumentIO Module Module1
Dim B2200 As New DirectIO("GPIB0::22::INSTR") Dim channels As String = "(@101,202,303,404)" '8 B2200.WriteLine("*RST") B2200.WriteLine(":ROUT:FUNC ACON") '10 B2200.WriteLine(":ROUT:CONN:RULE ALL,SROU") B2200.WriteLine(":ROUT:CONN:SEQ ALL,BBM") B2200.WriteLine(":ROUT:AGND:PORT ALL,12") '13 B2200.WriteLine(":ROUT:AGND:UNUSED ALL,'5,6,7,8'") B2200.WriteLine(":ROUT:AGND:CHAN:ENAB:CARD ALL") B2200.WriteLine(":ROUT:AGND:STAT ALL,ON") B2200.WriteLine(":ROUT:CLOS " & channels) '17 MsgBox("Click OK to start measurement.", vbOKOnly, "") Console.WriteLine("Measurement in progress. . ." & Chr(10)) 'insert the code for measurement
B2200.WriteLine(":ROUT:AGND:STAT ALL,OFF") '21 B2200.WriteLine(":ROUT:OPEN:CARD ALL") B2200.Close() MsgBox("Click OK to stop the program.", vbOKOnly, "") Console.WriteLine("Measurement completed." & Chr(10)) End Sub '26 End Module
8 Defines the channel list in the channels variable (string).
9 to 12 Resets the Agilent B2200. And sets the Auto configuration mode, the Single connection rule, and the Break_Before_Make connection sequence.
13 to 16 Sets the ground port and the ground enabled input/output ports. And turns the ground mode ON. Open the ground enabled input ports to prevent the instruments from damage.
17 Closes switches to connect the input-output paths specified by channels.
17 to 27 Almost same as the code shown in the lines 13 to 22 of Table 4-3. The command used to turn the ground mode OFF has been inserted to the line 21.
Programming Programming Examples · Sub Main() · Line Description

The following example uses the couple mode.
Figure 4-5 Couple Mode Example

from SMU1 to Output 1 (channel list 00101)
from SMU2 to Output 2 (channel list 00202)
from SMU3 to Output 3 (channel list 00303)
from SMU4 to Output 4 (coupled with the channel list 00303)
from SMU5 to Output 5 (channel list 00505)
from SMU6 to Output 6 (coupled with the channel list 00505)
Programming Programming Examples

Table 4-6 Couple Mode Example
Imports Agilent.TMFramework Imports Agilent.TMFramework.DataAnalysis Imports Agilent.TMFramework.DataVisualization Imports Agilent.TMFramework.InstrumentIO Module Module1
Dim B2200 As New DirectIO("GPIB0::22::INSTR") Dim channels As String = "(@101,202,303,505)" '8 B2200.WriteLine("*RST") B2200.WriteLine(":ROUT:FUNC ACON") '10 B2200.WriteLine(":ROUT:CONN:RULE ALL,SROU") B2200.WriteLine(":ROUT:CONN:SEQ ALL,BBM") B2200.WriteLine(":ROUT:COUP:PORT ALL,'3,5'") '13 B2200.WriteLine(":ROUT:COUP:STAT ALL,ON") B2200.WriteLine(":ROUT:CLOS " & channels) '15 MsgBox("Click OK to start measurement.", vbOKOnly, "") Console.WriteLine("Measurement in progress. . ." & Chr(10)) 'insert the code for measurement
B2200.WriteLine(":ROUT:COUP:STAT ALL,OFF") '19 B2200.WriteLine(":ROUT:OPEN:CARD ALL") B2200.Close() MsgBox("Click OK to stop the program.", vbOKOnly, "") Console.WriteLine("Measurement completed." & Chr(10)) End Sub '24 End Module
8 Defines the channel list in the channels variable (string).
9 to 12 Resets the Agilent B2200. And sets the Auto configuration mode, the Single connection rule, and the Break_Before_Make connection sequence.
13 to 14 Sets the couple ports, and turns the couple mode ON.
15 Closes switches to connect the input-output paths specified by channels.
15 to 25 Almost same as the code shown in the lines 13 to 22 of Table 4-3. The command used to turn the couple mode OFF has been inserted to the line 19.
Programming Programming Examples · Sub Main() · Line Description

You can define labels to use for the input/output ports when you control the switch in the GPIB local mode. The labels are cleared by *RST. So it is recommended to save the labels into the internal memory and define a comment for the memory data. You can see the comment when you load/save the memory data in the GPIB local mode.
Programming Programming Examples

Table 4-7 Label Definition and Data Save Example
Imports Agilent.TMFramework Imports Agilent.TMFramework.InstrumentIO Module Module1
Dim B2200 As New DirectIO("GPIB0::22::INSTR") B2200.WriteLine("*RST") B2200.WriteLine(":ROUT:FUNC ACON") Console.WriteLine("Starts labeling." & Chr(10)) B2200.WriteLine(":SYST:DISP:STR 'Updating memory 1 data.'") B2200.WriteLine(":ROUT:SYMB:PORT 1,'SMU1 '") '10 B2200.WriteLine(":ROUT:SYMB:PORT 2,'SMU2 '") B2200.WriteLine(":ROUT:SYMB:PORT 3,'SMU3 '") B2200.WriteLine(":ROUT:SYMB:PORT 4,'UNUSED'") B2200.WriteLine(":ROUT:SYMB:PORT 5,'SMU4-F'") B2200.WriteLine(":ROUT:SYMB:PORT 6,'SMU4-S'") B2200.WriteLine(":ROUT:SYMB:PORT 7,'SMU5-F'") B2200.WriteLine(":ROUT:SYMB:PORT 8,'SMU5-S'") B2200.WriteLine(":ROUT:SYMB:PORT 9,'OPEN '") B2200.WriteLine(":ROUT:SYMB:PORT 10,'BIAS '") B2200.WriteLine(":ROUT:SYMB:PORT 11,'OPEN '") B2200.WriteLine(":ROUT:SYMB:PORT 12,'GROUND'") B2200.WriteLine(":ROUT:SYMB:PORT 13,'CMU-H '") B2200.WriteLine(":ROUT:SYMB:PORT 14,'CMU-L '") B2200.WriteLine(":ROUT:SYMB:CHAN ALL,1,'BULK '") B2200.WriteLine(":ROUT:SYMB:CHAN ALL,2,'SOURCE'") B2200.WriteLine(":ROUT:SYMB:CHAN ALL,3,'GATE '") B2200.WriteLine(":ROUT:SYMB:CHAN ALL,4,'DRAIN '") B2200.WriteLine(":ROUT:SYMB:CHAN ALL,5,'GROUND'") '28 B2200.WriteLine(":SYST:MEMO:SAVE 1") B2200.WriteLine(":SYST:MEMO:COMM 1,'Port label info '") B2200.WriteLine(":SYST:DISP:STR 'Memory 1 data was updated.'") B2200.Close() Console.WriteLine("Labeling completed." & Chr(10)) End Sub '34 End Module
9, 31 Displays message on the LCD.
10 to 28 Defines labels for the inputs 1 to 14 and the outputs 1 to 5.
29 Saves the switch module setup including the labels into the internal memory 1.
30 Defines the comment “Port label info” for the memory 1.
Programming Programming Examples · Sub Main() · Line Description

You can define comments for the internal memories. You can see the comment when you load/save the memory data in the GPIB local mode.
Programming Programming Examples

Table 4-8 Memory Comment Definition Example
Imports Agilent.TMFramework Imports Agilent.TMFramework.DataAnalysis Imports Agilent.TMFramework.DataVisualization Imports Agilent.TMFramework.InstrumentIO Module Module1
Dim B2200 As New DirectIO("GPIB0::22::INSTR") B2200.WriteLine("*RST") '8 B2200.WriteLine(":ROUT:FUNC ACON") Console.WriteLine("Starts labeling." & Chr(10)) B2200.WriteLine(":SYST:DISP:STR 'Updating memory comment.'") B2200.WriteLine(":SYST:MEMO:COMM 1,'1-1,2-2,3-3,5-15'") '12 B2200.WriteLine(":SYST:MEMO:COMM 2,'1-1,2-2,3-4,5-17'") B2200.WriteLine(":SYST:MEMO:COMM 3,'1-1,2-2,3-5,5-19'") B2200.WriteLine(":SYST:MEMO:COMM 4,'1-1,2-2,5-7,7-21'") B2200.WriteLine(":SYST:MEMO:COMM 5,'1-1,2-2,5-9,7-23'") B2200.WriteLine(":SYST:MEMO:COMM 6,'10-BIAS '") B2200.WriteLine(":SYST:MEMO:COMM 7,'12-GROUND '") B2200.WriteLine(":SYST:MEMO:COMM 8,'CMH-10,CML-11 '") '19 B2200.WriteLine(":SYST:DISP:STR 'Memory comment was updated.'") B2200.Close() Console.WriteLine("Labeling completed." & Chr(10)) End Sub '23 End Module
8 Resets the Agilent B2200.
9 Sets the channel configuration mode to Auto.
12 to 19 Defines the comments for the internal memory 1 to 8.
Programming Programming Examples · Sub Main() · Line Description

When the capacitance/conductance measurement is performed through the Agilent B2200, LCR meter measures the capacitance/conductance of the path including a device under test (DUT), matrix switches, extension cables and so on. So, the data measured by the LCR meter is far from the DUT’s capacitance/conductance.
The Agilent B2200 VXIplug&play driver provides the functions used to compensate the capacitance/conductance measured by the Agilent 4284A LCR meter in the measurement environments described in “Required Conditions” on page 4-21.
This section explains how to use the capacitance compensation function.
Driver functions used for the capacitance compensation are listed below.
Capacitance measurement accuracy (typical): ±1 % ±0.5 pF
This typical data is for the following measurement conditions:
Measurement terminal: At the end of the Agilent 16494A/B/C cable connected to the switch module output terminals.
The typical data does not apply to anything extended from the 16494A/B/C cable. The conditions described in “Required Conditions” on page 4-21 must be satisfied.
Programming Capacitance Compensation

The following conditions must be satisfied to use the capacitance compensation function. For the instrument connections, see Figure 4-6.
Use the Agilent 16494F CMU cable or the Agilent 16048D/E test leads to connect between the Agilent 4284A and the Agilent B2200 inputs.
If the 16048D/E is used, the BNC-T adapters (2 ea., Agilent part number 1250-2405 for each) are required to connect between the Hc and Hp terminals and between the Lc and Lp terminals.
Perform the 4284A open calibration at the end of the measurement paths in front of the B2200 inputs. If you also perform the short calibration (optional), prepare the BNC thru adapter (Agilent part number 1250-0080, 1 ea.).
AUX Input 13 (CMH, for 4284A Hc-Hp) and 14 (CML, for 4284A Lc-Lp)
Use the Agilent 16494A triaxial cable or Agilent 16494B/C Kelvin triaxial cable.
You can also use another type of triaxial cable, coaxial cable, or combination of these.
To approximate the capacitance/conductance of the DUT, you need to obtain the appropriate compensation coefficients for your measurement environment, and create your compensation data file. See Figure 4-6.
Programming Capacitance Compensation

In Figure 4-6, C2H, C2L, C3H, C3L are the compensation coefficients defined in the compensation data file. where, CxH is for the path connected to the Agilent 4284A Hc-Hp terminal, and CxL is for the path connected to the Agilent 4284A Lc-Lp terminal.
When the Agilent B2220A probe card interface is used, obtain the coefficients for C3x, and create your compensation data file. In this case, probe card will be used for the C3x path.
When the connector plate is used, obtain the coefficients for C2x and C3x, and create your compensation data file. In this case, triaxial cable with connector plate will be used for the C2x path. And coaxial cable with positioner will be used for the C3x path.
For obtaining the compensation coefficients and creating the compensation data file, see “To Create Compensation Data File” on page 4-23.
Figure 4-6 Extension Cables and Compensation Coefficients

Programming Capacitance Compensation

This section explains how to create the compensation data file.
To select the most appropriate template for your measurement environment, see Table 4-9 that lists the file name and the measurement environment where the template targets. Each template is a text file that contains the information as shown below.
# # Compensation data file for B2210A/Probecard I/F/3m triax cable # # # MB Mother Board # MH Matrix Path High # ML Matrix Path Low # C1 Agilent Triax Cable # C2H Probe card I/F or User Triax cable High # C2L Probe card I/F or User Triax cable Low # C3H Probe card or User Coax cable High # C3L Probe card or User Coax cable Low # format version 1.0 B2210A PCIF # # R[Ohm] L[H] C[F] # MB 0.000000e+00 5.250000e-08 2.940000e-11 MH 2.430000e+00 6.310000e-07 1.930000e-10 ML 2.490000e+00 5.970000e-07 1.920000e-10 C1 6.300000e-01 1.250000e-06 1.600000e-10 C2H 2.988000e-01 5.090000e-07 7.000000e-11 C2L 2.988000e-01 5.090000e-07 7.000000e-11 C3H 0.000000e+00 8.000000e-08 1.500000e-13 C3L 0.000000e+00 8.000000e-08 1.500000e-13
The 15th line specifies the used switch module, B2210A or B2211A.
The 16th line specifies the DUT interface, PCIF or CABLE. PCIF indicates that the Agilent B2220A probe card interface is used. CABLE indicates that the connector plate is used.
The lines C2H to C3L should be modified for each measurement environment. See Table 4-9 and Table 4-10. Do not modify the other lines.
Programming Capacitance Compensation

Table 4-9 Template Compensation Data Files
Measurement environment that template targets
Switch module Cableb
\B2210A\pcif\triax\3m.data B2210A 16494A-002 B2220A C3H and C3L
\B2210A\pcif\triax\4m.data 16494A-005
\B2210A\pcif\kelvin\3m.data 16494C-002
\B2210A\pcif\kelvin\4m.data 16494C-005
\B2210A\cable\triax\1_5m.data 16494A-001 16495F/G C2H, C2L, C3H, and C3L \B2210A\cable\triax\3m.data 16494A-002
\B2210A\cable\triax\4m.data 16494A-005
\B2210A\cable\kelvin\1_5m.data 16494B-001
\B2210A\cable\kelvin\3m.data 16494B-002
\B2210A\cable\kelvin\4m.data 16494C-005
\B2211A\pcif\triax\3m.data B2211A 16494A-002 B2220A C3H and C3L
\B2211A\pcif\triax\4m.data 16494A-005
\B2211A\pcif\kelvin\3m.data 16494C-002
\B2211A\pcif\kelvin\4m.data 16494C-005
\B2211A\cable\triax\1_5m.data 16494A-001 16495F/G C2H, C2L, C3H, and C3L \B2211A\cable\triax\3m.data 16494A-002
\B2211A\cable\triax\4m.data 16494A-005
\B2211A\cable\kelvin\1_5m.data 16494B-001
\B2211A\cable\kelvin\3m.data 16494B-002
\B2211A\cable\kelvin\4m.data 16494C-005
b. Model number of the cable connected between the switch module and the DUT interface.
c. Agilent B2220A probe card interface or Agilent 16495F/G connector plate.

Table 4-10 Compensation Coefficients and Modifications
coefficients Modifications of data file
For the Agilent B2220A probe card interface, do not modify the lines.
For the connector plate, change the R, L, C values in the lines. The value must be changed to the R, L, C values of the C2x path (triaxial cable with connector plate) shown in Figure 4-6.
Change the R, L, C values in the lines. The value must be changed to the R, L, C values of the C3x path.
For the Agilent B2220A probe card interface, probe card will be used for the C3x path.
For the connector plate, coaxial cable with positioner will be used for the C3x path.
After the measurements, calculate the per meter value of the R, L, C, and record them into the following table.
C2H
C3H
Do not change any other lines. Also do not change the value for the coefficients that should not be modified.

Obtain the compensation coefficients as shown below.
a. Connect the path/cable corresponding to C2H to the Agilent 4284A. Then measure and record the R, L, and C values.
b. Connect the path/cable corresponding to C2L to the Agilent 4284A. Then
measure and record the R, L, and C values.
Table 4-11 R, L, C Measurement Conditions

a. Select 1 point. Do not change while measurements of all coefficients.
b. For triaxial cable, connect B to F directly, and measure L between A
and E. For coaxial cable, connect B to D directly, and measure L between A and C. Ignore E and F.
Figure 4-7 Measurement Terminals of C2H/C2L/C3H/C3L Path

Programming Capacitance Compensation

Perform the capacitance measurement and compensation as shown below.
See Table 4-13 for the example to use the capacitance compensation function. This example uses Microsoft Visual Basic .NET and Agilent T&M Programmers Toolkit. And the Agilent B2200 controlled by the program is defined as AGB2200 at the 10th line. So, the function names are changed as shown below.
agb220xa_selectCompenFile --> Agb2200.SelectCompenFile
agb220xa_compenC --> Agb2200.CompenC
Also, agb220xa_reset and agb220xa_close have been changed.
C = C2r - C1r
Table 4-12 Recording Measurement/Compensation Data
2 (measured) C1 = G1 =
3 (compensated) C1r = G1r =
4 (measured) C2 = G2 =
5 (compensated) C2r = G2r =
6 (calculated) C =
Programming Capacitance Compensation · Measurement/Compensation Data · step

Table 4-13 Capacitance Compensation Program Example
Imports Agilent.TMFramework Imports Agilent.TMFramework.DataAnalysis Imports Agilent.TMFramework.DataVisualization Imports Agilent.TMFramework.InstrumentIO Imports Agilent.TMFramework.InstrumentDriverInterop Imports Agilent.TMFramework.InstrumentDriverInterop.Design Imports Agilent.TMFramework.InstrumentDriverInterop.VxipnpWrappers Module Module1
Sub Main() ’9 Dim Agb2200 As Agb220xa = New Agb220xa("GPIB0::22::INSTR", True, True) Agb2200.Reset() Dim f_com as String = "C:\temp\my_env_1.txt" ’12 Agb2200.SelectCompenFile(f_com)
Dim freq As Double = 1000000 ’measurement frequency: 1 (MHz) 15 Dim data_c As Double = 0.0000000001 ’C measured by 4284A: 100 (pF) Dim data_g As Double = 0.0005 ’G measured by 4284A: 0.5 (mS) Dim res_c As Double Dim res_g As Double Agb2200.CompenC(freq, data_c, data_g, res_c, res_g) ’20
Dim result As String = "C = " & res_c * 1000000000000.0 & " (pF)" ’22 result = result & Chr(10) & "G = " & res_g * 1000 & " (mS)" & Chr(10) Console.WriteLine(result) Agb2200.Close() End Sub End Module
1 to 11 The above example is for the B2200 of the GPIB address 22 on the interface GPIB0. “GPIB0” is the VISA name. Confirm your GPIB settings, and set them properly.
12 to 13 The lines specify the compensation data file. The file name must specify your compensation data file.
15 to 20 Compensates the data measured by the Agilent 4284A. In this example, the measurement frequency is 1 MHz, the capacitance data is 100 pF, and the conductance data is 0.5 mS.
Change freq, data_c, data_g values for your measurement results.
22 to 24 Displays the compensation result data on the console window. Record the values as C1r and G1r, or C2r and G2r.
Programming Capacitance Compensation · Line Description


This chapter describes the following for Agilent B2200:
SCPI is a universal programming language for electronic test and measurement instruments, and is based on IEEE 488.1 and IEEE 488.2.
SCPI commands are divided into two types: common commands and subsystem commands.
Common commands are generally not measurement related, but are used to manage status registers, data storage, and so on. Common commands are defined by IEEE 488.2. All common commands begin with an asterisk, such as *RST.
Subsystem commands are mostly measurement related and some are general purpose. Subsystem commands have a hierarchical structure that uses a colon between mnemonics (keywords), such as :DIAG:TEST:CARD:CLE.
Commands for the following subsystems are available for the B2200:
DIAGnostic commands for performing self-test.
ROUTe commands for signal routing.
SYSTem commands for controlling other functions that are not related to instrument performance.
Commands are listed in alphabetical order in this chapter.
These are IEEE 488.2 status structures (registers), which can be set or read by the common commands.
SCPI Command Reference

Textual Notation for Subsystem Commands
CAPITAL LETTERS Capital letters are the minimally required letters of the command header. Lowercase letters are the long form (complete spelling), which you can omit if desired.
For example, for :SYSTem:CCONfig?, you only need to specify :SYST:CCON?.
[ ] Square brackets are used to enclose optional information not required for execution of the command sequence.
For example, for the [:ROUTe]:BIAS:PORT command, you can execute :ROUT:BIAS:PORT or :BIAS:PORT, which are exactly the same.
italic font Text you supply. For example, for card_number, you enter an allowed card number. Also, used to represent response data.
| Vertical bar can be read as “or” and is used to separate alternative parameter options.
{ } Braces (curly brackets) are used to enclose one or more parameters that may be included zero or more times.
< > Angular brackets indicate that the word or words enclosed represent something other than themselves. For example, .
SCPI Command Reference

SCPI Command Reference Commands Summary
This section describes common commands and queries, which are commands defined by IEEE 488.2.
The following table shows some common commands that are supported for the B2200.
*ESE(?) Standard Event Status Enable Command (Query)
*ESR? Standard Event Status Register Query
*OPC(?) Operation Complete Command (Query)
*SRE(?) Service Request Enable Command (Query)
*STB? Read Status Byte Query

*CLS
This command clears the Status Byte Register, the Standard Event Status Register, and the Error Queue. This command does not clear the enable registers. See “Status Reporting Structure” on page 5-53.
Also, this command stops the monitoring of pending operations by the *OPC command.
This command does not have query form.
This command sets the bits of the Standard Event Status “Enable” Register. 1 enables, 0 masks.
enable_number decimal integer (that is the sum of the binary-weighted values for the desired bits), hexadecimal, octal, or binary value
Query response enable_number
enable_number is decimal integer value that is the sum of the binary-weighted values of the “Enable” register bits.
Semantics The Standard Event Status “Enable” Register determines which bits of the Standard Event Status Register are enabled. Enabled bits are ORed together, and the result is reported to bit5 of the Status Byte Register.
The Standard Event Status “Enable” Register consists of 16 bits, but only the lower 8 bits are used, which correspond to the bits of the Standard Event Status Register. For details, see “Standard Event Status Enable Register” on page 5-59.

SCPI Command Reference *ESR?
The following table shows the bits of the Standard Event Status Register and binary-weighted decimal value of each bit.
0 1 OPC (Operation Complete)
1 2 not used
2 4 QYE (Query ERROR)
3 8 DDE (Device-Dependent ERROR)
4 16 EXE (Execution ERROR)
5 32 CME (Command ERROR)
6 64 not used
7 128 PON (Power on)
Example The following four lines enable the same bit (CME bit):
OUTPUT @Agb2200;"*ESE 32" using decimal numeric
OUTPUT @Agb2200;"*ESE #B100000" using binary numeric
OUTPUT @Agb2200;"*ESE #Q40" using octal numeric
OUTPUT @Agb2200;"*ESE #H20" using hexadecimal numeric
The following is example for query:
OUTPUT @Agb2200;"*ESE?" ENTER @Agb2200;A
This query command returns the present contents of the Standard Event Status Register.
Query response register
register decimal integer value that is the sum of the binary-weighted values for the set bits
bit binary-weight description · Syntax *ESR? · Parameter Explanation

Semantics The following table shows the bits of the Standard Event Status Register, and the binary-weighted decimal value of each bit.
0 1 OPC (Operation Complete)
1 2 not used. always 0.
2 4 QYE (Query ERROR)
3 8 DDE (Device-Dependent ERROR)
4 16 EXE (Execution ERROR)
5 32 CME (Command ERROR)
6 64 not used. always 0.
7 128 PON (Power on)
Example OUTPUT @Agb2200;"*ESR?" ENTER @Agb2200;A
This query command returns the ID of your B2200.
Query response AGILENT,model,0,revision
model character model number. B2200A or B2201A
revision character revision number. A.01.00 or later
Example DIM A$[50] OUTPUT @Agb2200;"*IDN?" ENTER @Agb2200;A$ PRINT A$
An example result of above program is:

SCPI Command Reference *OPC
This command starts to monitor pending operations, and sets/clears the Operation Complete (OPC) bit in the Standard Event Status Register as follows:
So, *OPC command is required to enable the OPC bit. To stop monitoring pending operations (disable OPC bit), execute the *CLS command.
For details about the Standard Event Status Register, see “Standard Event Status Register” on page 5-58. Also, see *WAI command.
Query response 1
*OPC? places ASCII character 1 into the Output Queue when all pending operations are completed. For details, see “Output Queue” on page 5-60.
The following example is for query:
OUTPUT @Agb2200;"*OPC?" ENTER @Agb2200;A
Syntax *OPC · Example OUTPUT @Agb2200;"*OPC"

*RST
This command performs an instrument reset. Status after *RST is shown below:
Bias-enabled Channels: All channels (output ports) are bias-enabled.
Couple Input Port: Cleared.
Ground-enabled Channels: Cleared.
Ground-enabled Ports: Cleared.
Card Channel Status: All relays are opened.
Self-Test Result: Not changed.
Input Port Symbol String: Cleared.
Output Port Symbol String: Cleared.
Light Pen: Enabled.
For the power-on settings, see :SYSTem:CPON.
SCPI Command Reference · Syntax *RST · Example OUTPUT @Agb2200;"*RST"

SCPI Command Reference *SRE
This command sets the Service Request “Enable” Register bits. 1 enables, 0 masks.
enable_number decimal integer (that is the sum of the binary-weighted values for the desired bits), hexadecimal, octal, or binary value
Query response enable_number
Semantics The Service Request “Enable” Register consists of 8 bits: Bit0 to Bit7. Bit6 is not defined, and is always 0. The Service Request “Enable” Register determines which bits of the Status Byte Register are enabled.
The status of the enabled bits are ORed together, and the result of OR is output to bit6 (Master Summary Status bit) of Status Byte Register. For details, see “Status Reporting Structure” on page 5-53.
The following table shows the bits of the Status Byte Register, and the binary-weighted decimal value of each bit.
0 1 not used
1 2 not used
2 4 not used
3 8 not used
4 16 MAV (Message Available summary-message)
5 32 ESB (Event Status Bit)
6 64 MSS (Master Summary Status)
7 128 not used
Example The following four lines enable the same bits (bit 4 and 5):
OUTPUT @Agb2200;"*SRE 48" using decimal numeric
OUTPUT @Agb2200;"*SRE #B110000" using binary numeric
OUTPUT @Agb2200;"*SRE #Q60" using octal numeric
OUTPUT @Agb2200;"*SRE #H30" using hexadecimal numeric
Syntax *SRE enable_number · Parameter Explanation · bit binary-weight description

The following is example for query:
OUTPUT @Agb2200;"*SRE?" ENTER @Agb2200;A
This query command reads the Status Byte Register (reads Master Summary Status bit, not Request for Service Message).
For bit6, this command reads MSS, not Request for Service (RQS). See “Status Reporting Structure” on page 5-53.
Query response register
register decimal integer value that is the sum of the binary-weighted values for the set bits
Semantics The following table shows the bits of the Status Byte Register, and the binary-weighted decimal value of each bit.
0 1 not used. always 0.
1 2 not used. always 0.
2 4 not used. always 0.
3 8 not used. always 0.
4 16 MAV (Message Available summary-message)
5 32 ESB (Event Status Bit)
6 64 MSS (Master Summary Status)
7 128 not used. always 0.
Example OUTPUT @Agb2200;"*STB?" ENTER @Agb2200;A

SCPI Command Reference *TST?
This query command executes an internal self-test, then returns the result. After this command execution, the B2200 becomes same status as after *RST command execution.
Query response test_result
Example OUTPUT @Agb2200;"*TST?" ENTER @Agb2200;A
This command stops execution of any commands until the Operation Complete (OPC) bit is set to 1, which means there is no pending operation. See *OPC command.

ROUT subsystem:ROUTe subsystem has commands for controlling the signal routing.
“Open a channel”: opens relays to disconnect the channel (that is, disconnects input port from output port).
“Close a channel”: closes relays to connect the channel (that is, connects input port to output port).
[:ROUT]:FUNC channel_config [:ROUT]:FUNC?
Sets the channel configuration mode. channel_config: ACON: Auto Config Mode NCON: Normal Config Mode
Query returns the present channel configuration: ACON or NCON.
[:ROUT]:CONN:RULE card_number,rule [:ROUT]:CONN:RULE? card_number
Sets the connection rule for the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config rule: FREE (free) or SROUte (single)
Query returns the connection rule of the specified card: FREE or SROU. card_number: Card to check. 0 for Auto Config, 1, 2, 3, or 4 for Normal Config.
[:ROUT]:CONN:SEQ card_number,sequence [:ROUT]:CONN:SEQ? card_number
Specifies the connection sequence mode for the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config sequence: NSEQ: No-Sequence mode BBM: Break-Before-Make mode MBBR: Make-Before-Break mode
Query returns the connections sequence mode of the specified card: NSEQ, BBM, or MBBR. card_number: Card to check. 0 for Auto Config, 1, 2, 3, or 4 for Normal Config.
SCPI Command Reference · Command Summary · Command Description

SCPI Command Reference Command Summary
[:ROUT]:SYMB:CHAN card_number,channel,'string' [:ROUT]:SYMB:CHAN? card_number,channel
Defines a string for the specified channel.
card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config
Query returns the symbol string set to the specified channel.
[:ROUT]:SYMB:PORT port,'string' [:ROUT]:SYMB:PORT? port
Defines a string for the specified input port.
Query returns the symbol string set to the specified input port.
[:ROUT]:OPEN:CARD card_number Disconnects all input ports from all output ports for the specified card.
card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config
[:ROUT]:OPEN[:LIST] (@channel_list) [:ROUT]:OPEN[:LIST]? (@channel_list)
Disconnects the input ports from output ports as specified in channel_list.
channel_list: Channels to open.
Query returns the status of the specified channels: 0 (closed) or 1 (opened). channel_list: Channels to check.
[:ROUT]:CLOS:CARD? card_number Returns channel_list of all closed channels for the specified card. “closed channel” means an input port connected to an output port.
card_number: Card to check. 0 for Auto Config, 1, 2, 3, or 4 for Normal Config.
[:ROUT]:CLOS[:LIST] (@channel_list) [:ROUT]:CLOS[:LIST]? (@channel_list)
Connects the input ports to the output ports as specified in channel_list.
channel_list: Channels to close.
Query returns the status of the specified channels: 1 (closed) or 0 (opened). channel_list: Channels to check.
Command Description · Relay Control Commands

[:ROUT]:BIAS:CHAN:DIS:CARD card_number Bias-disables the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config
[:ROUT]:BIAS:CHAN:DIS[:LIST] (@channel_list) [:ROUT]:BIAS:CHAN:DIS[:LIST]? (@channel_list)
Bias-disables the specified channels. channel_list: Channels to bias-disable.
Query returns the status of the specified channels: 1 (disabled) or 0 (enabled). channel_list: Channels to check.
[:ROUT]:BIAS:CHAN:ENAB:CARD card_number Bias-enables the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config
[:ROUT]:BIAS:CHAN:ENAB[:LIST] (@channel_list) [:ROUT]:BIAS:CHAN:ENAB[:LIST]? (@channel_list)
Bias-enables the specified channels. channel_list: Channels to bias-enable.
Query returns the status of the specified channels: 1 (enabled) or 0 (disabled). channel_list: Channels to check.
[:ROUT]:BIAS:PORT card_number,bias_port [:ROUT]:BIAS:PORT? card_number
Specifies the input Bias Port for the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config bias_port: 1 to 14 or -1
Query returns the input Bias Port number for the specified card. card_number: Card to check. 0 for Auto Config, 1, 2, 3, or 4 for Normal Config.
[:ROUT]:BIAS[:STAT] card_number,state [:ROUT]:BIAS[:STAT]? card_number
Sets the bias mode for the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config state: ON / 1 (mode ON) or OFF / 0 (mode OFF)
Query returns the mode status of the specified card: 0 (OFF) or 1 (ON). card_number: Card to check. 0 for Auto Config, 1, 2, 3, or 4 for Normal Config.

SCPI Command Reference Command Summary
[:ROUT]:AGND:CHAN:DIS:CARD card_number Ground-disables the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config
[:ROUT]:AGND:CHAN:DIS[:LIST] (@channel_list) [:ROUT]:AGND:CHAN:DIS[:LIST]? (@channel_list)
Ground-disables the specified channels. channel_list: Channels to ground-disable.
Query returns the status of the specified channels: 1 (disabled) or 0 (enabled). channel_list: Channels to check.
[:ROUT]:AGND:CHAN:ENAB:CARD card_number Ground-enables the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config
[:ROUT]:AGND:CHAN:ENAB[:LIST] (@channel_list) [:ROUT]:AGND:CHAN:ENAB[:LIST]? (@channel_list)
Ground-enables the specified channels. channel_list: Channels to ground-enable.
Query returns the status of the specified channels: 1 (enabled) or 0 (disabled). channel_list: Channels to check.
[:ROUT]:AGND:PORT card_number,ground_port [:ROUT]:AGND:PORT? card_number
Specifies the input Ground Port for the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config ground_port: 1 to 14 or -1
Query returns the input Ground Port number for the specified card. card_number: Card to check. 0 for Auto Config, 1, 2, 3, or 4 for Normal Config.
[:ROUT]:AGND[:STAT] card_number,state [:ROUT]:AGND[:STAT]? card_number
Sets the ground mode for the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config state: ON / 1 (mode ON) or OFF / 0 (mode OFF)
Query returns the mode status of the specified card: 0 (OFF) or 1 (ON). card_number: Card to check. 0 for Auto Config, 1, 2, 3, or 4 for Normal Config.
Command Description · Ground Mode Commands

[:ROUT]:AGND:UNUSED card_number,'enable_port' [:ROUT]:AGND:UNUSED? card_number
Ground-enables the specified input ports for the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config enable_port: One or more input port numbers: 1 to 8. Enclose by single quotation marks. Separate multiple input port numbers by comma. For example: '1,5'
Query returns the ground-enabled input port numbers for the specified card. card_number: Card to check. 0 for Auto Config, 1, 2, 3, or 4 for Normal Config.
[:ROUT]:COUP:PORT card_number,'couple_port' [:ROUT]:COUP:PORT? card_number
Specifies the input couple ports for the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config couple_port: One or more input port numbers: 1, 3, 5, 7, 9, 11, or 13. Enclose by single quotation marks. Separate multiple input port numbers by comma. For example: '1,5'
Query returns the lower input port number of each couple pair for the specified card. card_number: Card to check. 0 for Auto Config, 1, 2, 3, or 4 for Normal Config.
[:ROUT]:COUP:PORT:DET Detects the input ports connected to Kelvin cable, and assigns them as the input couple ports for the all cards.
[:ROUT]:COUP[:STAT] card_number,state [:ROUT]:COUP[:STAT]? card_number
Sets the couple mode for the specified card. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config state: ON / 1 (mode ON) or OFF / 0 (mode OFF)
Query returns the mode status of the specified card: 0 (OFF) or 1 (ON). card_number: Card to check. 0 for Auto Config, 1, 2, 3, or 4 for Normal Config.

SCPI Command Reference Command Summary
DIAG subsystem:DIAGnostic subsystem has commands for executing the self-test function. For more info, see “Selftest Menu” on page 3-34.
The :DIAGnostic subsystem commands ignore the B2200 channel configuration mode. For :DIAG commands that require a card number, you specify 1, 2, 3, 4, or ALL.
:DIAG:TEST:CARD:CLE card_number Clears relay test result (pass/fail) of the specified card.
:DIAG:TEST:CARD[:EXEC]? card_number Executes relay test, then returns result: 1 (fail card exists), 0 (pass).
:DIAG:TEST:CARD:STAT? card_number Returns most recent relay test result: 1 (fail), 0 (pass), -1 (not tested).
:DIAG:TEST:FRAM:CLE item Clears specified test result.
:DIAG:TEST:FRAM[:EXEC]? item Executes specified test, then returns test result: 1 (fail), 0 (pass).
:DIAG:TEST:FRAM:STAT? item Returns most recent test result of the specified test: 1 (fail), 0 (pass), -1 (not tested).
Command Description

:SYSTem subsystem is a collection of functions that are not related to instrument performance.
:SYST:BEEP state Enables/disables the beeper. state: ON / 1 (enable) or OFF / 0 (disable)
:SYST:CCON? card_number Returns the card configuration information. This command is just to keep compatibility with the Agilent E5250A. card_number: 1, 2, 3, or 4
:SYST:CDES? card_number Returns a description of the specified card: model number and input/output port information. card_number: Card to check. 0 for Auto Config, 1, 2, 3, or 4 for Normal Config.
:SYST:CPON card_number Resets the specified card to the power-on state. card_number: 0 or ALL for Auto Config, 1, 2, 3, 4, or ALL for Normal Config
:SYST:CTYP? card_number Returns ID of the specified card: model number and revision. card_number: Card to check. 0 for Auto Config, 1, 2, 3, or 4 for Normal Config.
:SYST:DISP:LCD state Enables/disables the front panel LCD when the B2200 is in the GPIB remote mode. state: ON / 1 (enable) or OFF / 0 (disable)
:SYST:DISP:LED state Enables/disables the front panel LED. state: ON / 1 (enable) or OFF / 0 (disable)
:SYST:DISP:STR string Specifies a string displayed on the LCD in the GPIB remote mode.
:SYST:ERR? Reads error from head of error queue, and removes it from the queue.
:SYST:KLC state Locks/unlocks the front panel keys. state: ON / 1 (lock) or OFF / 0 (unlock)
:SYST:MEMO:SAVE memory_number:SYST:MEMO:LOAD memory_number
Saves a setup information into the internal memory, or loads a setup information. memory_number: 1 to 8

SCPI Command Reference Command Summary
:SYST:MEMO:COMM memory_number,'comment' :SYST:MEMO:COMM? memory_number
Memorizes the comment for the B2200 setup information specified by memory_number. memory_number: 1 to 8
:SYST:MEMO:DEL memory_number Deletes the B2200 setup information and the comment specified by memory_number. memory_number: 1 to 8
:SYST:PEN state Enables/disables the light pen. state: ON / 1 (enable) or OFF / 0 (disable)
:SYST:VERS? Returns SCPI version number for which the B2200 complies.
Command Description
![Page 147 of the Agilent Technologies B2201A user manual — :DIAGnostic:TEST:CARD[:EXECute]?](/cdn/reader/4e/4e82065aba2cdae36fe10b22abbd9da21332896df8d5b1360dd3c4b19da908c8/images/page-147.webp?w=400)
SCPI Command Reference :DIAGnostic:TEST:CARD:CLEar
This command clears the relay test result (pass/fail result) of the specified card.
Syntax:DIAGnostic:TEST:CARD:CLEar card_number
Example OUTPUT @Agb2200;":DIAG:TEST:CARD:CLE 1"
This command executes the relay test for specified card, then returns the pass/fail result. Before starting the relay test, open the input/output terminals (end of cable is OK, Kelvin cable must be removed). For more information, see “RELAY_TEST” on page 3-35.
Syntax :DIAGnostic:TEST:CARD[:EXECute]? card_number
Query response test_result
Example OUTPUT @Agb2200;":DIAG:TEST:CARD? ALL"
Parameter Explanation · Parameter Explanation

SCPI Command Reference :DIAGnostic:TEST:CARD:STATe?
This command returns the most recent relay test result for the specified card.
Syntax :DIAGnostic:TEST:CARD:STATe? card_number
Query response test_result
Example OUTPUT @Agb2200;":DIAG:TEST:CARD:STAT? 1" ENTER @Agb2200;A
This command clears test result of the specified B2200 test.
Syntax :DIAGnostic:TEST:FRAMe:CLEar CONTroller | FPANel | LED | PEN | BEEPer
FPANel Front panel interface test
LED LED matrix test
PEN Light pen test
Example OUTPUT @Agb2200;":DIAG:TEST:FRAM:CLE CONT"
![Page 149 of the Agilent Technologies B2201A user manual — :DIAGnostic:TEST:FRAMe[:EXECute]?](/cdn/reader/4e/4e82065aba2cdae36fe10b22abbd9da21332896df8d5b1360dd3c4b19da908c8/images/page-149.webp?w=400)
SCPI Command Reference :DIAGnostic:TEST:FRAMe[:EXECute]?
This command executes the specified B2200 test, then returns the test result. See “Selftest Menu” on page 3-34 to perform the test.
After the controller test, the B2200 status becomes same as after *RST command execution.
Syntax :DIAGnostic:TEST:FRAMe[:EXECute]? CONTroller | FPANel | LED | PEN | BEEPer
FPANel Front panel interface test
LED LED matrix test
PEN Light pen test
Query response test_result
Example OUTPUT @Agb2200;":DIAG:TEST:FRAM? CONT"
Parameter Explanation · CONTroller Controller test · BEEPer Beeper test
![Page 150 of the Agilent Technologies B2201A user manual — [:ROUTe]:AGND:CHANnel:DISable:CARD](/cdn/reader/4e/4e82065aba2cdae36fe10b22abbd9da21332896df8d5b1360dd3c4b19da908c8/images/page-150.webp?w=400)
SCPI Command Reference :DIAGnostic:TEST:FRAMe:STATe?
This command returns the most recent test result (pass/fail) of the specified test.
Syntax :DIAGnostic:TEST:FRAMe:STATe? CONTroller | FPANel | LED | PEN | BEEPer
FPANel Front panel interface test
LED LED matrix test
PEN Light pen test
Query response test_result
Example OUTPUT @Agb2200;":DIAG:TEST:FRAM:STAT? CONT" ENTER @Agb2200;A
This command ground-disables the all output ports (channels) for the specified card. When the Ground Mode is ON, the ground-disabled output ports are disconnected from the input Ground Port. At *RST, no channel is ground-enabled. The ground mode is set by “[:ROUTe]:AGND[:STATe]”.
Syntax [:ROUTe]:AGND:CHANnel:DISable:CARD card_number
card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL
Example OUTPUT @Agb2200;":ROUT:AGND:CHAN:DIS:CARD ALL"
![Page 151 of the Agilent Technologies B2201A user manual — [:ROUTe]:AGND:CHANnel:ENABle:CARD](/cdn/reader/4e/4e82065aba2cdae36fe10b22abbd9da21332896df8d5b1360dd3c4b19da908c8/images/page-151.webp?w=400)
SCPI Command Reference [:ROUTe]:AGND:CHANnel:DISable[:LIST]
This command specifies the ground-disabled output ports (channels). When the Ground Mode is ON, the ground-disabled output ports are disconnected from the input Ground Port. At *RST, no channel is ground-enabled. The ground mode is set by “[:ROUTe]:AGND[:STATe]”.
The query returns whether the specified channels are ground-disabled or not.
Syntax [:ROUTe]:AGND:CHANnel:DISable[:LIST] (@channel_list)
[:ROUTe]:AGND:CHANnel:DISable[:LIST]? (@channel_list)
channel_list Channels to ground-disable or to check. For channel_list, see “Switch Control” on page 4-5.
Input port is always the input Ground Port. So, the input ports in channel_list are ignored. However, you cannot abbreviate the input port.
Query response disable_status{, disable_status}
Example OUTPUT @Agb2200;":ROUT:AGND:CHAN:DIS (@10101)" OUTPUT @Agb2200;":ROUT:AGND:CHAN:DIS? (@10101,10102,10201)" ENTER @Agb2200;A$
This example ground-disables the output port 1 on the card 1. In this example, A$ will be 1,0,1.
This command ground-enables the all output ports (channels) for the specified card. When the Ground Mode is ON, the ground-enabled output ports that have not been connected to any other input port are connected to the input Ground Port. This command is available only for the specified card. At *RST, no channel is ground-enabled. The ground mode is set by “[:ROUTe]:AGND[:STATe]”.
Parameter Explanation
![Page 152 of the Agilent Technologies B2201A user manual — [:ROUTe]:AGND:CHANnel:ENABle[:LIST]](/cdn/reader/4e/4e82065aba2cdae36fe10b22abbd9da21332896df8d5b1360dd3c4b19da908c8/images/page-152.webp?w=400)
SCPI Command Reference [:ROUTe]:AGND:CHANnel:ENABle[:LIST]
Syntax [:ROUTe]:AGND:CHANnel:ENABle:CARD card_number
card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL
Example OUTPUT @Agb2200;":ROUT:AGND:CHAN:ENAB:CARD ALL"
This command specifies the ground-enabled output ports (channels). When the Ground Mode is ON, the ground-enabled output ports that have not been connected to any other input port are connected to the input Ground Port. Then, the input Ground Port will not be connected to the output ports that have been connected to any other input port. At *RST, no channel is ground-enabled. The ground mode is set by “[:ROUTe]:AGND[:STATe]”.
The query returns whether the specified channels are ground-enabled or not.
Syntax [:ROUTe]:AGND:CHANnel:ENABle[:LIST] (@channel_list)
[:ROUTe]:AGND:CHANnel:ENABle[:LIST]? (@channel_list)
channel_list Channels to ground-enable or to check. For channel_list, see “Switch Control” on page 4-5.
Input port is always the input Ground Port. So, the input ports in channel_list are ignored. However, you cannot abbreviate the input port number.
Query response enable_status{, enable_status}
Example OUTPUT @Agb2200;":ROUT:AGND:CHAN:ENAB (@10101)" OUTPUT @Agb2200;":ROUT:AGND:CHAN:ENAB? (@10101,10102,10201)" ENTER @Agb2200;A$
This example ground-enables the output port 1 on the card 1. In this example, A$ will be 1,0,1.
Parameter Explanation · Parameter Explanation
![Page 153 of the Agilent Technologies B2201A user manual — [:ROUTe]:AGND:PORT](/cdn/reader/4e/4e82065aba2cdae36fe10b22abbd9da21332896df8d5b1360dd3c4b19da908c8/images/page-153.webp?w=400)
[:ROUTe]:AGND:PORT
This command specifies the input Ground Port for the specified card. For each card, you can specify the same or different Ground Port. At *RST, the Ground Port is 12. The ground mode is set by “[:ROUTe]:AGND[:STATe]”.
The query returns the input port number of the Ground Port.
If the input ground port and a couple port have been assigned to the same input port, the ground mode and the couple mode cannot be used in parallel.
0 V output voltage source should be connected to the input Ground Port. Or set ground_port to 12, and open the input port 12 that is the default ground port. If you assign the input 12 as the ground port, this ground port will be internally connected to the ground when the ground mode is set to ON.
Syntax [:ROUTe]:AGND:PORT card_number,ground_port
[:ROUTe]:AGND:PORT? card_number
card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL ALL is not available for query.
ground_port Input port number: 1 to 14. Or -1 to disable ground port.
Query response port_number
port_number = 1 to 14 or -1. Number set to ground_port.
Example OUTPUT @Agb2200;":ROUT:AGND:PORT 0,13" OUTPUT @Agb2200;":ROUT:AGND:PORT? 0" ENTER @Agb2200;A
In this example, A will be 13.
SCPI Command Reference · Parameter Explanation
![Page 154 of the Agilent Technologies B2201A user manual — [:ROUTe]:AGND[:STATe]](/cdn/reader/4e/4e82065aba2cdae36fe10b22abbd9da21332896df8d5b1360dd3c4b19da908c8/images/page-154.webp?w=400)
SCPI Command Reference [:ROUTe]:AGND[:STATe]
This command controls the Ground Mode for the specified card. When the Ground Mode is ON, the input Ground Port is connected to the all ground-enabled input ports/output ports that have not been connected to any other port. At *RST, the Ground Mode is OFF.
The query returns the mode status.
The input Ground Port is set by “[:ROUTe]:AGND:PORT”. The ground-enabled input ports are set by “[:ROUTe]:AGND:UNUSED”. And the ground-enabled output ports are set by “[:ROUTe]:AGND:CHANnel:ENABle[:LIST]” or “[:ROUTe]:AGND:CHANnel:ENABle:CARD”.
When the Ground Mode is ON, you cannot directly control the Ground Port connections. However, they can be indirectly controlled when the connections of the other input port are controlled.
When the Ground Mode is OFF, the input Ground Port is the same as the other input ports, so relays can be controlled directly to connect to output ports.
Syntax [:ROUTe]:AGND[:STATe] card_number,state
[:ROUTe]:AGND[:STATe]? card_number
card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL ALL is not available for query.
Query response state
Example OUTPUT @Agb2200;":ROUT:AGND:STAT 0,ON" OUTPUT @Agb2200;":ROUT:AGND:STAT? 0" ENTER @Agb2200;A
In this example, A will be 1.
Parameter Explanation
![Page 155 of the Agilent Technologies B2201A user manual — [:ROUTe]:AGND:UNUSED](/cdn/reader/4e/4e82065aba2cdae36fe10b22abbd9da21332896df8d5b1360dd3c4b19da908c8/images/page-155.webp?w=400)
[:ROUTe]:AGND:UNUSED
This command specifies the ground-enabled input ports for the specified card. When the Ground Mode is ON, the ground-enabled input ports that have not been connected to any other port are connected to the input Ground Port. At *RST, no input port is ground-enabled. The ground mode is set by “[:ROUTe]:AGND[:STATe]”.
The query returns which input ports are ground-enabled.
If the ground enabled input port and a couple port have been assigned to the same input port, the ground mode and the couple mode cannot be used in parallel.
Syntax [:ROUTe]:AGND:UNUSED card_number,'port_number'
[:ROUTe]:AGND:UNUSED? card_number
card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL ALL is not available for query.
port_number Input port number: 1 to 8.
Enclose by single quotation marks. Separate multiple input port numbers by comma. For example: '1,5'
Query response port_number{, port_number}
port_number = 1 to 8
Example OUTPUT @Agb2200;":ROUT:AGND:UNUSED 0,'5,6,7,8'" OUTPUT @Agb2200;":ROUT:AGND:UNUSED? 0" ENTER @Agb2200;A$
In this example, A$ will be 5,6,7,8.
SCPI Command Reference · Parameter Explanation
![Page 156 of the Agilent Technologies B2201A user manual — [:ROUTe]:BIAS:CHANnel:DISable[:LIST]](/cdn/reader/4e/4e82065aba2cdae36fe10b22abbd9da21332896df8d5b1360dd3c4b19da908c8/images/page-156.webp?w=400)
SCPI Command Reference [:ROUTe]:BIAS:CHANnel:DISable:CARD
This command bias-disables the all output ports (channels) for the specified card. When the Bias Mode is ON, the bias-disabled output ports are disconnected from the input Bias Port. At *RST, all cards are bias-enabled. The bias mode is set by “[:ROUTe]:BIAS[:STATe]”.
Syntax [:ROUTe]:BIAS:CHANnel:DISable:CARD card_number
card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL
Example OUTPUT @Agb2200;":ROUT:BIAS:CHAN:DIS:CARD ALL"
This command specifies the bias-disabled output ports (channels). When the Bias Mode is ON, the bias-disabled output ports are disconnected from the input Bias Port. At *RST, all channels are bias-enabled. The bias mode is set by “[:ROUTe]:BIAS[:STATe]”.
The query returns whether the specified channels are bias-disabled or not.
Syntax [:ROUTe]:BIAS:CHANnel:DISable[:LIST] (@channel_list)
[:ROUTe]:BIAS:CHANnel:DISable[:LIST]? (@channel_list)
channel_list Channels to bias-disable or to check. For channel_list, see “Switch Control” on page 4-5.
Input port is always the input Bias Port. So, the input ports in channel_list are ignored. However, you cannot abbreviate the input port.
Query response disable_status{, disable_status}
Parameter Explanation · Parameter Explanation
![Page 157 of the Agilent Technologies B2201A user manual — [:ROUTe]:BIAS:CHANnel:ENABle[:LIST]](/cdn/reader/4e/4e82065aba2cdae36fe10b22abbd9da21332896df8d5b1360dd3c4b19da908c8/images/page-157.webp?w=400)
SCPI Command Reference [:ROUTe]:BIAS:CHANnel:ENABle:CARD
Example OUTPUT @Agb2200;":ROUT:BIAS:CHAN:ENAB:CARD ALL" OUTPUT @Agb2200;":ROUT:BIAS:CHAN:DIS (@10101)" OUTPUT @Agb2200;":ROUT:BIAS:CHAN:DIS? (@10101,10102,10201)" ENTER @Agb2200;A$
This example bias-disables the output port 1 on the card 1. In this example, A$ will be 1,0,1.
This command bias-enables the all output ports (channels) for the specified card. When the Bias Mode is ON, the bias-enabled output ports that have not been connected to any other input port are connected to the input Bias Port. This command is available only for the specified card. At *RST, all cards are bias-enabled. The bias mode is set by “[:ROUTe]:BIAS[:STATe]”.
Syntax [:ROUTe]:BIAS:CHANnel:ENABle:CARD card_number
card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL
Example OUTPUT @Agb2200;":ROUT:BIAS:CHAN:ENAB:CARD ALL"
This command specifies the bias-enabled output ports (channels). When the Bias Mode is ON, the bias-enabled output ports that have not been connected to any other input port are connected to the input Bias Port. Then, the input Bias Port will not be connected to the output ports that have been connected to any other input port. At *RST, all channels are bias-enabled. The bias mode is set by “[:ROUTe]:BIAS[:STATe]”.
The query returns whether the specified channels are bias-enabled or not.
Syntax [:ROUTe]:BIAS:CHANnel:ENABle[:LIST] (@channel_list)
[:ROUTe]:BIAS:CHANnel:ENABle[:LIST]? (@channel_list)
channel_list Channels to bias-enable or to check. For channel_list, see “Switch Control” on page 4-5.
Parameter Explanation · Parameter Explanation
![Page 158 of the Agilent Technologies B2201A user manual — [:ROUTe]:BIAS:PORT](/cdn/reader/4e/4e82065aba2cdae36fe10b22abbd9da21332896df8d5b1360dd3c4b19da908c8/images/page-158.webp?w=400)
SCPI Command Reference [:ROUTe]:BIAS:PORT
Input port is always the input Bias Port. So, the input ports in channel_list are ignored. However, you cannot abbreviate the input port.
Query response enable_status{, enable_status}
Example OUTPUT @Agb2200;":ROUT:BIAS:CHAN:DIS:CARD ALL" OUTPUT @Agb2200;":ROUT:BIAS:CHAN:ENAB (@10101)" OUTPUT @Agb2200;":ROUT:BIAS:CHAN:ENAB? (@10101,10102,10201)" ENTER @Agb2200;A$
This example bias enables the output port 1 on the card 1. In this example, A$ will be 1,0,1.
This command specifies the input Bias Port for the specified card. For each card, you can specify the same or different Bias Port. At *RST, Bias Port is set to 10. The bias mode is set by “[:ROUTe]:BIAS[:STATe]”. The query returns the input Bias Port number for the specified card.
Syntax [:ROUTe]:BIAS:PORT card_number,bias_port
[:ROUTe]:BIAS:PORT? card_number
card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL ALL is not available for query.
bias_port Input port number: 1 to 14. Or -1 to disable bias port.
Query response port_number
port_number = 1 to 14 or -1. Number set to bias_port.
Example OUTPUT @Agb2200;":ROUT:BIAS:PORT ALL,4" OUTPUT @Agb2200;":ROUT:BIAS:PORT? 1" ENTER @Agb2200;A
Parameter Explanation

[:ROUTe]:BIAS[:STATe]
In this example, A will be 4.
This command controls the Bias Mode for the specified card. When the Bias Mode is ON, the input Bias Port is connected to the all bias-enabled output ports that have not been connected to any other input port. At *RST, the Bias Mode is OFF.
The query returns the mode status of the specified card.
The input Bias Port is set by “[:ROUTe]:BIAS:PORT”. The bias-enabled output ports (channels) are set by “[:ROUTe]:BIAS:CHANnel:ENABle[:LIST]” or “[:ROUTe]:BIAS:CHANnel:ENABle:CARD”.
When the Bias Mode is ON, you cannot directly control the Bias Port connections. However, they can be indirectly controlled when the connections of the other input port are controlled.
When the Bias Mode is OFF, the input Bias Port is the same as the other input ports, so relays can be controlled directly to connect to output ports.
Syntax [:ROUTe]:BIAS[:STATe] card_number,state
[:ROUTe]:BIAS[:STATe]? card_number
card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL ALL is not available for query.
Query response 0 or 1
Example OUTPUT @Agb2200;":ROUT:BIAS:STAT ALL,ON" OUTPUT @Agb2200;":ROUT:BIAS:STAT? 4" ENTER @Agb2200;A
In this example, A will be 1.
SCPI Command Reference · Parameter Explanation
![Page 160 of the Agilent Technologies B2201A user manual — [:ROUTe]:CLOSe[:LIST]](/cdn/reader/4e/4e82065aba2cdae36fe10b22abbd9da21332896df8d5b1360dd3c4b19da908c8/images/page-160.webp?w=400)
SCPI Command Reference [:ROUTe]:CLOSe:CARD?
This query command returns channel_list of all closed (connected) channels for the specified card.
Syntax [:ROUTe]:CLOSe:CARD? card_number
Query response channel_list
Example OUTPUT @Agb2200;":ROUT:OPEN:CARD ALL" OUTPUT @Agb2200;":ROUT:CLOS (@10101,10202)" OUTPUT @Agb2200;":ROUT:CLOS:CARD? 1" ENTER @Agb2200;A$
A$ will be @10101,10202. For channel_list, see “Switch Control” on page 4-5.
This command connects the input ports to the output ports (channels) as specified in channel_list. The query returns 0 or 1 for each channel specified by channel_list, in same order as specified by channel_list.
For some modes (such as Single Route Mode, Couple Mode, Ground Mode, and Bias Mode), more complex connections may occur.
Syntax [:ROUTe]:CLOSe[:LIST] (@channel_list)
[:ROUTe]:CLOSe[:LIST]? (@channel_list)
channel_list Channels to close or to check. For channel_list, see “Switch Control” on page 4-5.
Query response close_status{, close_status}
Parameter Explanation · Parameter Explanation
![Page 161 of the Agilent Technologies B2201A user manual — [:ROUTe]:CONNection:RULE](/cdn/reader/4e/4e82065aba2cdae36fe10b22abbd9da21332896df8d5b1360dd3c4b19da908c8/images/page-161.webp?w=400)
SCPI Command Reference [:ROUTe]:CONNection:RULE
Example OUTPUT @Agb2200;":ROUT:CLOS (@10101,10202)" OUTPUT @Agb2200;":ROUT:CLOS? (@10101,10102,10201,10202)" ENTER @Agb2200;A$
In this example, A$ will be 1,0,0,1.
This command sets the connection rule (Free or Single Route) for the specified card. The query returns the connection rule of the specified card. At *RST, all cards are set to FREE. See “Connection Rule” on page 3-14.
Syntax [:ROUTe]:CONNection:RULE card_number,rule
[:ROUTe]:CONNection:RULE? card_number
card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL ALL is not available for query.
rule FREE (free route) or SROUte (single route).
In the free route, multiple channel connections are available for all input ports.
In the single route, only 1 channel connection is available for each input port.
Query response FREE or SROU
Example OUTPUT @Agb2200;":ROUT:CONN:RULE ALL,SROU" OUTPUT @Agb2200;":ROUT:CONN:RULE? 1" ENTER @Agb2200;A$
In this example, A$ will be SROU.
Parameter Explanation
![Page 162 of the Agilent Technologies B2201A user manual — [:ROUTe]:CONNection:SEQuence](/cdn/reader/4e/4e82065aba2cdae36fe10b22abbd9da21332896df8d5b1360dd3c4b19da908c8/images/page-162.webp?w=400)
SCPI Command Reference [:ROUTe]:CONNection:SEQuence
This command specifies the connection sequence mode, which is the open/close sequence of the relays when connection route is changed from an existing connection to a new connection. The query returns the connection sequence mode of the specified card. At *RST, BBMake is selected. See “Connection Sequence” on page 3-15.
The connection sequence is used only for cards that have been set to SROUte connection rule. Also see “[:ROUTe]:CONNection:RULE”.
Syntax [:ROUTe]:CONNection:SEQuence card_number,sequence
[:ROUTe]:CONNection:SEQuence? card_number
card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL ALL is not available for query.
Query response NSEQ or BBM or MBBR
Example OUTPUT @Agb2200;":ROUT:CONN:SEQ ALL,MBBR" OUTPUT @Agb2200;":ROUT:CONN:SEQ? 0" ENTER @Agb2200;A$
In this example, A$ will be MBBR.
Parameter Explanation
![Page 163 of the Agilent Technologies B2201A user manual — [:ROUTe]:COUPle:PORT](/cdn/reader/4e/4e82065aba2cdae36fe10b22abbd9da21332896df8d5b1360dd3c4b19da908c8/images/page-163.webp?w=400)
[:ROUTe]:COUPle:PORT
If the input ground port and a couple port have been assigned to the same input port, the ground mode and the couple mode cannot be used in parallel.
If the ground enabled input port and a couple port have been assigned to the same input port, the ground mode and the couple mode cannot be used in parallel.
This command is used to set up input couple ports, which are used for the Kelvin connections. The couple mode is set by “[:ROUTe]:COUPle[:STATe]”.
For each card, you can set up the same or different input couple ports.
This command overwrites the previous couple port setting. The couple port setting will be cleared and updated by “[:ROUTe]:COUPle:PORT:DETect”.
The query returns the odd input port number of each coupled pair.
At *RST, no input ports are coupled.
Syntax [:ROUTe]:COUPle:PORT card_number,'couple_port'
[:ROUTe]:COUPle:PORT? card_number
card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL ALL is not available for query.
couple_port One or more input port numbers: 1, 3, 5, 7, 9, 11 or 13.
Enclose by single quotation marks. Separate multiple input port numbers by comma. For example: '1,5'
Query response port_number{, port_number}
port_number = 1 or 3 or 5 or 7 or 9 or 11 or 13
Example OUTPUT @Agb2200;":ROUT:COUP:PORT ALL,'1,3'" OUTPUT @Agb2200;":ROUT:COUP:PORT? 1" ENTER @Agb2200;A$
In this example, A$ will be 1,3.
SCPI Command Reference · Parameter Explanation
![Page 164 of the Agilent Technologies B2201A user manual — [:ROUTe]:COUPle:PORT:DETect](/cdn/reader/4e/4e82065aba2cdae36fe10b22abbd9da21332896df8d5b1360dd3c4b19da908c8/images/page-164.webp?w=400)
SCPI Command Reference [:ROUTe]:COUPle:PORT:DETect
This command detects the input ports connected to the Kelvin cable, and assigns them as the input couple ports that will be used for the Kelvin connection. The input couple port setting is effective for the all cards. This command overwrites the previous couple port setting.
The couple mode is set by “[:ROUTe]:COUPle[:STATe]”. The couple port setting will be cleared and updated by “[:ROUTe]:COUPle:PORT”.
The input port numbers of the couple ports can be got by [:ROUTe]:COUPle:PORT?.
After this command is executed, the all relay settings will be the same as after the :ROUT:OPEN:CARD ALL command execution.
Syntax [:ROUTe]:COUPle:PORT:DETect
Example OUTPUT @Agb2200;":ROUT:COUP:PORT:DET"

[:ROUTe]:COUPle[:STATe]
This command controls the Couple Mode for the specified card. The query returns the mode status of the specified card. At *RST, the Couple Mode is OFF.
The input couple ports are set by “[:ROUTe]:COUPle:PORT:DETect” or “[:ROUTe]:COUPle:PORT”.
If you specify a connection from a couple input port to an output port when the couple mode is ON, the B2200 automatically controls relays to connect the input ports n-1 and n to the output ports m-1 and m respectively (n: an even number from 2 to 14, m: an even number from 2 to 12, 2 to 24, 2 to 36, or 2 to 48 depends on the B2200 configuration). See example below.
Syntax [:ROUTe]:COUPle[:STATe] card_number,state
[:ROUTe]:COUPle[:STATe]? card_number
card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL ALL is not available for query.
Query response state
Example OUTPUT @Agb2200;":ROUT:COUP:STAT ALL,ON" OUTPUT @Agb2200;":ROUT:COUP:STAT? 2" ENTER @Agb2200;A
In this example, A will be 1.
OUTPUT @Agb2200;":ROUT:COUP:PORT ALL,'1'" OUTPUT @Agb2200;":ROUT:COUP:STAT ALL,ON" OUTPUT @Agb2200;":ROUT:CLOS (@10103)"
This example controls the card 1 relays and connects the inputs 1 and 2 to the outputs 3 and 4 respectively. :ROUT:CLOS (@10104) will make the same result.
SCPI Command Reference · Parameter Explanation
![Page 166 of the Agilent Technologies B2201A user manual — [:ROUTe]:OPEN:CARD](/cdn/reader/4e/4e82065aba2cdae36fe10b22abbd9da21332896df8d5b1360dd3c4b19da908c8/images/page-166.webp?w=400)
SCPI Command Reference [:ROUTe]:FUNCtion
This command sets the channel configuration, which determines how you specify the channel_list or card_number in other commands. The query returns the present channel configuration. At *RST, this parameter is set to ACONfig. See “Channel Configuration Mode” on page 3-13.
When configuration is changed by this command, all channels are opened (disconnected) and become same status as after :SYSTem:CPON ALL execution.
Syntax [:ROUTe]:FUNCtion channel_config
[:ROUTe]:FUNCtion?
Query response ACON or NCON
Example OUTPUT @Agb2200;":ROUT:FUNC ACON" OUTPUT @Agb2200;":ROUT:FUNC?" ENTER @Agb2200;A$
In this example, A$ will be ACON.
This command disconnects all input ports from all output ports for the specified card. For some modes (such as Single Route Mode, Couple Mode, Ground Mode, and Bias Mode), more complex disconnections may occur.
Syntax [:ROUTe]:OPEN:CARD card_number
card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL
Example OUTPUT @Agb2200;":ROUT:OPEN:CARD ALL"
Parameter Explanation · Parameter Explanation
![Page 167 of the Agilent Technologies B2201A user manual — [:ROUTe]:OPEN[:LIST]](/cdn/reader/4e/4e82065aba2cdae36fe10b22abbd9da21332896df8d5b1360dd3c4b19da908c8/images/page-167.webp?w=400)
[:ROUTe]:OPEN[:LIST]
This command disconnects the input ports from output ports as specified in channel_list. The query returns 0 or 1 for each channel specified by channel_list, in same order as specified by channel_list.
For some modes (such as Single Route Mode, Couple Mode, Ground Mode, and Bias Mode), more complex disconnections may occur.
Syntax [:ROUTe]:OPEN[:LIST] (@channel_list)
[:ROUTe]:OPEN[:LIST]? (@channel_list)
channel_list Channels to open (disconnect) or to check. For channel_list, see “Switch Control” on page 4-5.
Query response open_status{, open_status}
Example OUTPUT @Agb2200;":ROUT:OPEN:CARD ALL" OUTPUT @Agb2200;":ROUT:CLOS (@10101,10202)" OUTPUT @Agb2200;":ROUT:OPEN? (@10101,10102,10201,10202)" ENTER @Agb2200;A$
In this example, A$ will be 0,1,1,0.
SCPI Command Reference · Parameter Explanation
![Page 168 of the Agilent Technologies B2201A user manual — [:ROUTe]:SYMBol:CHANnel](/cdn/reader/4e/4e82065aba2cdae36fe10b22abbd9da21332896df8d5b1360dd3c4b19da908c8/images/page-168.webp?w=400)
SCPI Command Reference [:ROUTe]:SYMBol:CHANnel
In the GPIB local mode, the B2200 uses 01 to 48 (two digits numbers) to specify the output ports (channels) 1 to 48 respectively. This command specifies a symbol string and lets the B2200 use the specified string instead of the two digits number. The query returns the symbol string.
Maximum value of the two digits number depends on the B2200 configuration. It is 12 for the normal configuration mode or the auto configuration mode with one matrix card, 24 for the auto configuration mode with two matrix cards, 36 for the auto configuration mode with three matrix cards, or 48 for the auto configuration mode with four matrix cards.
Reboot, *RST, and :SYSTem:CPON clear the symbol strings. After that, the initial symbol string (two digits numbers 01 to 48) will be set.
Syntax [:ROUTe]:SYMBol:CHANnel card_number,channel_number,'string'
[:ROUTe]:SYMBol:CHANnel? card_number,channel_number
card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL ALL is not available for query.
channel_number Output port number: 1 to 48 for Auto, 1 to 12 for Normal.
string Symbol string set to the specified output port (channel). Maximum 6 characters.
Numeric characters, alphabetic characters; upper case and lower case, space, and the following symbols are available.
! " # $ % & ’ ( ) * + , - . / > ? : ~ ; < = > ? @ [ \ ] ‘ { | } ^ _
Query response string
Example OUTPUT @Agb2200;":ROUT:SYMB:CHAN 0,10,'GATE'" OUTPUT @Agb2200;":ROUT:SYMB:CHAN? 0,10" ENTER @Agb2200;A$
This example assigns the string GATE to the output port 10. After that, GATE is used instead of 10. In this example, A$ will be GATE.
Parameter Explanation
![Page 169 of the Agilent Technologies B2201A user manual — [:ROUTe]:SYMBol:PORT](/cdn/reader/4e/4e82065aba2cdae36fe10b22abbd9da21332896df8d5b1360dd3c4b19da908c8/images/page-169.webp?w=400)
[:ROUTe]:SYMBol:PORT
In the GPIB local mode, the B2200 uses 01 to 14 (two digits numbers) to specify the input ports 1 to 14 respectively. This command specifies a symbol string and lets the B2200 use the specified string instead of the two digits number. The query returns the symbol string.
Reboot and *RST clear the symbol strings. After that, the initial symbol string (two digits numbers 01 to 14) will be set.
:SYSTem:CPON does not clear the symbol strings.
Syntax [:ROUTe]:SYMBol:PORT port_number,'symbol_string'
[:ROUTe]:SYMBol:PORT? port_number
port_number Input port number: 1 to 14.
symbol_string Symbol string set to the specified input port. Maximum 6 characters.
Numeric characters, alphabetic characters; upper case and lower case, space, and the following symbols are available.
! " # $ % & ’ ( ) * + , - . / > ? : ~ ; < = > ? @ [ \ ] ‘ { | } ^ _
Query response symbol_string
Example OUTPUT @Agb2200;":ROUT:SYMB:PORT 1,'SMU1'" OUTPUT @Agb2200;":ROUT:SYMB:PORT? 1" ENTER @Agb2200;A$
This example assigns the string SMU1 to the input port 1. After that, SMU1 is used instead of 01. In this example, A$ will be SMU1.
SCPI Command Reference · Parameter Explanation

SCPI Command Reference :SYSTem:BEEP
This command enables/disables the beeper.
Syntax :SYSTem:BEEP state
Example OUTPUT @Agb2200;":SYST:BEEP ON"
This query command is just to keep compatibility with the Agilent E5250A Low Leakage Switch Mainframe. This query command returns the card configuration information.
Syntax :SYSTem:CCONfig? card_number
Query response card_configuration
Always returns #10.
Example OUTPUT @Agb2200;":SYST:CCON? 1" ENTER @Agb2200;A$ PRINT "Card configuration = ";A$
Parameter Explanation · Parameter Explanation

:SYSTem:CDEScription?
This query command returns a description of the specified card.
Syntax :SYSTem:CDEScription? card_number
Query response "card_description"
If the designated card is not installed or initialize failed, this command returns "No Card".
For B2200A:
"B2210A 14x12 Femto Leakage Switch Module"
For B2201A:
"B2211A 14x12 Low Leakage Switch Module"
Example OUTPUT @Agb2200;":SYST:CDES? 1" ENTER @Agb2200;A$
SCPI Command Reference · Parameter Explanation

SCPI Command Reference :SYSTem:CPON
This command resets the specified card. Status after this command is shown below:
Channel Configuration: Not changed.
Bias-enabled Channels: All channels (output ports) are bias-enabled.
Couple Input Port: Cleared.
Ground-enabled Channels: Cleared.
Ground-enabled Ports: Cleared.
Card Channel Status: All relays are opened.
Self-Test Result: Not changed.
Input Port Symbol String: Not changed.
Output Port Symbol String: Cleared.
For the device reset state, see *RST.
Syntax :SYSTem:CPON card_number
card_number For Auto Config mode: 0 or ALL For Normal Config mode: 1, 2, 3, 4, or ALL
Example OUTPUT @Agb2200;":SYST:CPON ALL"
Parameter Explanation

:SYSTem:CTYPe?
This query command returns the ID of the specified card.
Syntax :SYSTem:CTYPe? card_number
Query response AGILENT TECHNOLOGIES,model,0,revision
model = B2210A or B2211A
revision = 1 or later version
If designated card is not installed or initialize failed, this command returns NONE,NONE,0,0.
Example OUTPUT @Agb2200;":SYST:CTYPE? 1" ENTER @Agb2200;A$
This command enables/disables the front panel Liquid Crystal Display when the B2200 is in the GPIB remote mode. The front panel LCD will be automatically set to ON by the mode transition from GPIB remote to GPIB local.
Syntax :SYSTem:DISPlay:LCD state
Even though state is OFF, some status indicators on the LCD will be effective.
Example OUTPUT @Agb2200;":SYST:DISP:LCD 1"
SCPI Command Reference · Parameter Explanation · Parameter Explanation

SCPI Command Reference :SYSTem:DISPlay:LED
This command enables/disables the front panel Light Emitting Diodes matrix when the B2200 is in the GPIB remote mode. The front panel LED matrix will be automatically set to ON by the mode transition from GPIB remote to GPIB local.
Syntax :SYSTem:DISPlay:LED state
Example OUTPUT @Agb2200;":SYST:DISP:LED 1"
When the B2200 is in the GPIB remote mode, you can display a string (maximum 39 characters) on the front panel LCD. This command specifies the string and displays it on the LCD.
The specified string data is only displayed on the front panel LCD. It is not memorized. So a display-change operation will clear and delete the string data.
Syntax :SYSTem:DISPlay:STRing 'string'
string String to be displayed on the front panel LCD. Maximum 38 characters.
Numeric characters, alphabetic characters; upper case and lower case, space, and the following symbols are available.
! " # $ % & ’ ( ) * + , - . / > ? : ~ ; < = > ? @ [ \ ] ‘ { | } ^ _
Example OUTPUT @Agb2200;":SYST:DISP:STR 'E:SMU1,B:SMU2,C:SMU3'"
This example displays E:SMU1,B:SMU2,C:SMU3 on the front panel LCD.
Parameter Explanation · Parameter Explanation

:SYSTem:ERRor?
This query command reads the error from the head of the error queue and removes that error from the queue.
Syntax :SYSTem:ERRor?
Query response err_no,"message"
err_no is numeric response data, and message is string response data.
If there has been no error (error queue is empty), the response to this query is as follows:
Example OUTPUT @Agb2200;":SYST:ERR?" ENTER @Agb2200;A,B$
This command locks/unlocks the front panel keys.
Syntax :SYSTem:KLC state
Example OUTPUT @Agb2200;":SYST:KLC ON"
SCPI Command Reference · 0,"No error" · Parameter Explanation

SCPI Command Reference :SYSTem:MEMOry:COMMent
This command memorizes the comment for the B2200 setup information specified by memory_number. The previous comment will be deleted. The query returns the comment for the specified setup.
Syntax :SYSTem:MEMOry:COMMent memory_number,'comment'
:SYSTem:MEMOry:COMMent? memory_number
memory_number memory number: 1 to 8.
comment Comment. Maximum 16 characters. If this command is entered with empty comment, the comment will be cleared.
Numeric characters, alphabetic characters; upper case and lower case, space, and the following symbols are available.
! " # $ % & ’ ( ) * + , - . / > ? : ~ ; < = > ? @ [ \ ] ‘ { | } ^ _
Example OUTPUT @Agb2200;":SYST:MEMO:COMM 1,'1-1,2-13,3-25'" OUTPUT @Agb2200;":SYST:MEMO:COMM? 1" ENTER @Agb2200;A$
In this example, A$ will be 1-1,2-13,3-25.
This command deletes the B2200 setup information and the comment specified by memory_number.
Syntax :SYSTem:MEMOry:DELete memory_number
memory_number memory number: 1 to 8.
Example OUTPUT @Agb2200;":SYST:MEMO:DEL 1"
Parameter Explanation · Parameter Explanation

:SYSTem:MEMOry:LOAD
This command restores the B2200 setup information specified by memory_number. It must be already saved by “:SYSTem:MEMOry:SAVE”.
Syntax :SYSTem:MEMOry:LOAD memory_number
memory_number memory number: 1 to 8.
Example OUTPUT @Agb2200;":SYST:MEMO:SAVE 1" OUTPUT @Agb2200;":SYST:MEMO:LOAD 1"
This command saves the present setup information of the B2200 into the internal memory specified by memory_number. The previous setup will be deleted. To memorize the comment for the setup information, use “:SYSTem:MEMOry:COMMent”.
The following setup information will be saved.
Syntax :SYSTem:MEMOry:SAVE memory_number
memory_number memory number: 1 to 8.
Example OUTPUT @Agb2200;":SYST:MEMO:SAVE 1"
SCPI Command Reference · Parameter Explanation · Parameter Explanation

SCPI Command Reference :SYSTem:PEN
This command enables/disables the light pen.
Syntax :SYSTem:PEN state
Example OUTPUT @Agb2200;":SYST:PEN ON"
This query command returns the SCPI version number for which the B2200 complies.
Syntax :SYSTem:VERSion?
Query response YYYY.V
YYYY is the year (for example, 2004).
V is the approved revision number for that year.
Example OUTPUT @Agb2200;":SYST:VERS?" ENTER @Agb2200;A$
Parameter Explanation

This section describes the status reporting structure used in the B2200. These are IEEE 488.2 status structures that can be set and read by the SCPI Common Commands as described in “Common Commands” on page 5-4.
The status reporting structure consists of the following:
In general, the status reporting structure is used to request service via Service Request (SRQ) or to indicate a specific condition, such as operation complete.
See the following figures for a model of the B2200 status reporting structure.
SCPI Command Reference · Status Reporting Structure

SCPI Command Reference Status Reporting Structure
Figure 5-1 Status Reporting Structure of B2200

The above figure shows Service Request Generation. When a condition occurs that requires service, the instrument sets Request Service bit (RQS − bit6) of Status Byte, and sends a Service Request (SRQ) via GPIB bus to the controller. So, the controller can execute an interrupt service routine (must be in the program) that uses a Serial Poll to read Status Byte of each instrument to determine which instrument requested service. See “Status Byte Register” on page 5-55.

The B2200 Status Byte Register contains bits (ESB and MAV) for status summary messages from other registers. The status of these bits depends on the condition of the Standard Event Status Register and the Output Queue. If service request occurs, Bit6 (RQS) of Status Byte Register is set.
See Figure 5-2 and Table 5-1 for the B2200 Status Byte Register.
Figure 5-2 Status Byte Register of B2200

The Status Byte Register can be read with either a serial poll or the *STB? common query command (see “Common Commands” on page 5-4).
Serial poll is a low-level GPIB command that can be executed by SPOLL in HP BASIC, such as the following:
Both serial poll and STB? read the Status Byte Register identically, but have following difference:
RQS and MSS are always the same value, so returned Status Byte value is always the same for these two methods.
SCPI Command Reference · Status Byte Register · Status=SPOLL(@Agb2200)

SCPI Command Reference Status Byte Register
In general, use serial polling (not *STB?) inside interrupt service routines. Use *STB? in other cases (not in interrupt service routine) when you want to know the value of Status Byte.
Table 5-1 Status Byte Register of B2200
0 Not Used. Always zero.
1 Not Used. Always zero.
2 Not Used. Always zero.
3 Not Used. Always zero.
4 Message Available (MAV) Queue Summary Message
Indicates whether Output Queue is empty. If unread message exists, this bit is set to 1. Otherwise, this bit is set to 0. See “Output Queue” on page 5-60.
5 Standard Event Status Bit (ESB) Summary Message
Indicates whether one or more of the enabled Standard Events Status Register bits is set.
6 Request Service (RQS) Message
Indicates whether a SRQ (Service Request) has occurred. Read by Serial Poll.
Master Summary Status (MSS) Summary Message
Indicates that the instrument has at least one reason for requesting service. Read by *STB?.
7 Not Used. Always zero.
Bit Definition Explanation

SCPI Command Reference Service Request Enable Register
The Service Request Enable Register is an 8-bit register that can be used by the programmer to select which summary messages in the Status Byte Register may cause service requests. See Figure 5-3.
Figure 5-3 Service Request Enable Register of B2200


SCPI Command Reference Standard Event Status Register
The Standard Event Status Register has specific events assigned to specific bits. See Figure 5-4 and Table 5-2.
Figure 5-4 Standard Event Status Register of B2200


SCPI Command Reference Standard Event Status Enable Register
Table 5-2 Standard Event Status Register of B2200
0 Operation Complete (OPC)
This event bit has meaning only if a request to monitor is set by the *OPC command. See “*OPC” on page 5-8. This bit is set to 1 if there are no pending operations.
1 Not Used Always 0.
2 Query Error (QYE) • An attempt is being made to read data from the Output Queue when no data is present or pending.
3 Device Dependent Error (DDE)
This event bit indicates that an error has occurred which is not a Command Error, a Query Error, or an Execution Error.
4 Execution Error (EXE) Syntax of command is correct, but cannot be executed due to some condition of the B2200.
5 Command Error (CME)
A command syntax error has been detected.
6 Not Used Always 0.
7 Power On (PON) This event bit indicates that an off-to-on transition has occurred in instrument's power supply.
8 to 15 Not used Always 0.
The Standard Event Status “Enable” Register is an 8-bit register that can be used by the programmer to select which bits of Standard Event Status Register are enabled. The status of the enabled bits are ORed together, and result of OR will be reported to the ESB bit (Bit5) of the Status Byte Register.
The 8 bits of this register correspond to the 8 bits of the Standard Event Status Register. See Figure 5-4.
Bit Definition Explanation

SCPI Command Reference Output Queue
The Output Queue stores response messages until they are read. If an unread message exists, Bit4 (Message Available − MAV) of the Status Byte Register is set to 1. So, Bit4 is used to synchronize information exchange with the controller. See Figure 5-5.
Figure 5-5 Output Queue of B2200

The Response Formatter places Data Byte Messages and END messages into the Output Queue in response to query commands. These messages are removed from the Output Queue as they are read by the controller. As long as the Output Queue contains an unread message, MAV is 1.


This chapter introduces the VXI plug&play driver available for Agilent B2200, and consists of the following sections:
VXI plug&play Driver

VXI plug&play Driver System Requirements
The following system environments are required.
Microsoft Windows XP Professional, Windows 2000, Windows NT 4.0, or Windows 95. It must be supported by the application development environment.
Microsoft Visual Basic, Microsoft Visual C++, Borland C++Builder, National Instruments LabWindows or LabVIEW, or Agilent VEE.
Agilent W1140A or equivalent. Needed for Visual Basic .NET users.
Agilent 82357A USB/GPIB interface, E5810A LAN/GPIB gateway, 82350B GPIB interface, or equivalent. These models include Agilent VISA and SICL I/O libraries.
Required specifications depend on the application development environment. See manual of the application development environment. The flexible disk drive is required to install the VXIplug&play driver. The CD-ROM drive is required to install the software needed to use driver.
1 MB for the Agilent B2200 VXIplug&play driver

VXI plug&play Driver Installing VXIplug&play Driver
The installation flow is shown below. If you have already installed the GPIB interface card, VISA I/O library, and programming software on your PC, skip steps 1 through 5.
See the interface card manual. Note the model number of the interface card, as you may need it to configure the interface (in step 3).
Follow the instructions in the I/O library’s setup program.
See the I/O library manual.
Follow the setup program instructions.
a. Insert the Agilent B2200 Plug&Play Driver Disk into the flexible disk drive connected to your PC.
b. Execute the SETUP.EXE program on the diskette. The program
automatically installs the following files.

Table 6-1 lists all the functions for the Agilent B2200. You will see a brief description of the functions in the table.
For the description, syntax and parameters of the function, refer to the reference section following this table. The driver functions in the reference section will appear in the alphabetical order.
Table 6-1 Agilent B2200 Driver Function Lists
Miscellaneous agb220xa_init Initializes the B2200. agb220xa_close Closes the connection with the B2200. agb220xa_reset Executes the B2200 reset. agb220xa_self_test Executes the B2200 self-test. agb220xa_error_query Queries for the B2200 error code/message. agb220xa_error_message Queries for the driver error. agb220xa_revision_query Queries for the B2200 firmware/driver revisions. agb220xa_timeOut Sets the timeout. agb220xa_timeOut_Q Queries for the timeout setting. agb220xa_errorQueryDetect Sets the automatic error checking. agb220xa_errorQueryDetect_Q Queries for the automatic error checking setting. agb220xa_dcl Sends the Device Clear. agb220xa_esr_Q Queries for the ESR status. agb220xa_readStatusByte_Q Reads the B2200 status byte. agb220xa_opc_Q Checks the B2200 operation completion status. Mode Control agb220xa_func Sets the configuration mode. agb220xa_connRuleSeq Sets the connection rule/sequence. Couple Mode agb220xa_couplePort Selects the couple ports of the specified card. agb220xa_coupleState Sets the couple mode ON or OFF for the specified card. agb220xa_detectCouplePort Detects and sets the couple ports for all cards. Bias Mode agb220xa_biasPort Selects the input bias port of the specified card. agb220xa_biasChanCard Bias-enables all output ports of the specified card. agb220xa_biasChanList Bias-enables all output ports of the specified channel list. agb220xa_biasState Sets the bias mode ON or OFF for the specified card. agb220xa_biasChanList_Q Queries for the bias channel list.
VXI plug&play Driver · Driver Functions · Category Function Description

VXI plug&play Driver Driver Functions
Ground Mode agb220xa_groundPort Selects the input ground port of the specified card. agb220xa_groundChanCard Ground-enables all output ports of the specified card. agb220xa_groundChanList Ground-enables all output ports of the specified channel list. agb220xa_unusedPort Ground-enables the specified input ports of the specified card. agb220xa_groundState Sets the ground mode ON or OFF for the specified card. agb220xa_groundChanList_Q Queries for the ground channel list. Route Control agb220xa_closeList Closes the channel list. agb220xa_openList Opens the channel list. agb220xa_openCard Opens all output on the card. agb220xa_closeList_Q Queries for the channel list status. agb220xa_openList_Q agb220xa_closeCard_Q Queries for the closed channel list on the card. C/G Compensation
agb220xa_compenC Performs the compensation for the C/G data measured by the Agilent 4284A. agb220xa_selectCompenFile Selects the compensation data file used by the agb220xa_compenC function. Diagnostics agb220xa_testExec_Q Performs diagnostics and returns the test result. agb220xa_testClear Clears the specified test result. Passthrough Functions
agb220xa_cmd Sends a command. agb220xa_cmdInt Sends a command with an integer parameter. agb220xa_cmdReal Sends a command with a real parameter. agb220xa_cmdData_Q Sends a command to read any data. agb220xa_cmdString_Q Sends a command to read string response. agb220xa_cmdInt16_Q Sends a command to read 16 bit integer response. agb220xa_cmdInt16Arr_Q Sends a command to read 16 bit integer array response. agb220xa_cmdInt32_Q Sends a command to read 32 bit integer response. agb220xa_cmdInt32Arr_Q Sends a command to read 32 bit integer array response. agb220xa_cmdReal64_Q Sends a command to read 64 bit real response. agb220xa_cmdReal64Arr_Q Sends a command to read 64 bit real array response.
Category Function Description

This function bias-enables or disables all the output ports of the specified card.
Syntax ViStatus _VI_FUNC agb220xa_biasChanCard(ViSession vi, ViInt16 disable_enable, ViInt16 bias_cardno);
Parameters vi Instrument handle returned from agb220xa_init( ).
disable_enable Status of the card, bias enabled or bias disabled.
bias_cardno Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5 (all card) in the normal configuration mode, or 0 (all card in the auto configuration mode). For the configuration mode, see agb220xa_func.
This function bias-enables or disables all the output ports specified by the biaschan_list.
The parameter “biaschan_list” is an array of integers with each integer representing one channel. The last number of the “biaschan_list” should be “0” (numeric zero) to identify the end of the list. The maximum number of channels that can be specified by the list is 100.
Syntax ViStatus _VI_FUNC agb220xa_biasChanList(ViSession vi, ViInt16 biaschan_disen, ViInt32_VI_FAR biaschan_list[ ] );
Parameters vi Instrument handle returned from agb220xa_init( ).
biaschan_disen Status of the port, bias enabled or bias disabled.
biaschan_list[ ] Channel numbers. 5 digits integer. ABCDE. where A: card number, BC: input port number, DE: output port number. Top zero(s) can be ignored. For example, if A=0, BC=01, and DE=01, channel number should be 101 instead of 00101.
VXI plug&play Driver · Driver Functions

VXI plug&play Driver Driver Functions
This function will query the instrument for the status, bias enabled or bias disabled, for the channels given in the list.
The parameter “biaschan_list” is an array of integers with each integer representing one channel. The last number of the “biaschan_list” should be “0” (numeric zero) to identify the end of the list. The maximum number of channels that can be specified by the list is 100.
The “bias_status” parameter is an array of integers containing the return values of the query. The “bias_status” array returned will correspond one to one with “biaschan_list” parameter.
Syntax ViStatus _VI_FUNC agb220xa_biasChanList_Q(ViSession vi, ViInt16 bias_disen, ViInt32_VI_FAR biaschan_list[ ], ViInt32_VI_FAR bias_status[ ] );
Parameters vi Instrument handle returned from agb220xa_init( ).
bias_disen Query type.
biaschan_list[ ] Channel numbers to check the status. 5 digits integer. ABCDE. where A: card number, BC: input port number, DE: output port number. Top zero(s) can be ignored. For example, if A=0, BC=01, and DE=01, channel number should be 101 instead of 00101.
bias_status[ ] Status of the channels given in the biaschan_list. Returned value depends on the setting of bias_disen as shown below:
when bias_disen=0, 0 means bias disabled, 1 means enabled.
when bias_disen=1, 0 means bias enabled, 1 means disabled.

This function will select which input port is the bias port on the specified card. For each card, you can specify the same or different bias port.
Syntax ViStatus _VI_FUNC agb220xa_biasPort(ViSession vi, ViInt16 bport_cardno, ViInt16 bias_port);
Parameters vi Instrument handle returned from agb220xa_init( ).
bport_cardno Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5 (all card) in the normal configuration mode, or 0 (all card in the auto configuration mode). For the configuration mode, see agb220xa_func.
bias_port Input port number to be set to the bias port. 1 to 14 (input port 1 to input port 14).
If the bias input port and a couple port have been assigned to the same input port, the bias mode and the couple mode cannot be used in parallel.
This function controls the bias mode for the specified card. When bias mode is ON, the input bias port is connected to all bias enabled output ports that are not connected to any other input ports. Bias disabled output ports are never connected to the input bias port when bias mode is ON.
The bias mode cannot be set to ON when the ground mode is ON.
Syntax ViStatus _VI_FUNC agb220xa_biasState(ViSession vi, ViInt16 bstate_cardno, ViInt16 state);
Parameters vi Instrument handle returned from agb220xa_init( ).
bstate_cardno Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5 (all card) in the normal configuration mode, or 0 (all card in the auto configuration mode). For the configuration mode, see agb220xa_func.
state Bias mode. 0 (OFF) or 1 (ON).
VXI plug&play Driver · Driver Functions

VXI plug&play Driver Driver Functions
This function terminates the software connection to the instrument and deallocates system resources. It is generally a good programming habit to close the instrument handle when the program is done using the instrument.
Syntax ViStatus _VI_FUNC agb220xa_close(ViSession vi);
Parameters vi Instrument handle returned from agb220xa_init( ).
This function will query the card for the channels closed of the specified card.
The parameter “closechan_list” contains the channel numbers returned by the instrument. This will be an array of integers terminated by “zero” to identify the end of the list. Array of enough length should be passed to the function.
Syntax ViStatus _VI_FUNC agb220xa_closeCard_Q(ViSession vi, ViInt16 close_card, ViInt32_VI_FAR closechan_list[ ] );
Parameters vi Instrument handle returned from agb220xa_init( ).
close_card Card number. 1 (card 1), 2 (card 2), 3 (card 3), or 4 (card 4) in the normal configuration mode, or 0 (all card in the auto configuration mode). For the configuration mode, see agb220xa_func.
closechan_list[ ] Channels closed of the specified card.

This function will connect the input ports to the output ports specified by the channel list.
The parameter “closechan_list” is an array of integers with each integer representing one channel. The last number of the “closechan_list” should be “0” (numeric zero) to identify the end of the list. The maximum number of channels that can be specified by the list is 100.
Syntax ViStatus _VI_FUNC agb220xa_closeList(ViSession vi, ViInt32_VI_FAR closechan_list[ ] );
Parameters vi Instrument handle returned from agb220xa_init( ).
closechan_list[ ] Channel numbers to connect. 5 digits integer. ABCDE. where A: card number, BC: input port number, DE: output port number. Top zero(s) can be ignored. For example, if A=0, BC=01, and DE=01, channel number should be 101 instead of 00101.
VXI plug&play Driver · Driver Functions

VXI plug&play Driver Driver Functions
This function will query the instrument for the channels closed given in the “closechan_list”.
The parameter “closechan_list” is an array of integers with each integer representing one channel. The last number of the “closechan_list” should be “0” (numeric zero) to identify the end of the list. The maximum number of channels that can be specified by the list is 100.
The “close_status” parameter is an array of integers containing the return values of the query. The “close_status” array returned will correspond one to one with “closechan_list” parameter.
Syntax ViStatus _VI_FUNC agb220xa_closeList_Q(ViSession vi, ViInt32_VI_FAR closechan_list[ ], ViInt32_VI_FAR close_status[ ] );
Parameters vi Instrument handle returned from agb220xa_init( ).
closechan_list[ ] Channel numbers to know the close status. 5 digits integer. ABCDE. where A: card number, BC: input port number, DE: output port number. Top zero(s) can be ignored. For example, if A=0, BC=01, and DE=01, channel number should be 101 instead of 00101.
close_status[ ] Status of the channels given in the closechan_list. 0 (opened) or 1 (closed).
This function passes the cmd_str string to the instrument. Must be a NULL terminated C string.
Syntax ViStatus _VI_FUNC agb220xa_cmd(ViSession vi, ViString cmd_str);
Parameters vi Instrument handle returned from agb220xa_init( ).
cmd_str Instrument command (cannot exceed 256 bytes in length).

This function passes the cmd_str string to the instrument. This entry point will wait for a response which may be any data. You specify the cmd_str and size parameters, and get result[ ].
Syntax ViStatus _VI_FUNC agb220xa_cmdData_Q(ViSession vi, ViString cmd_str, ViInt32 size, ViChar_VI_FAR result[ ] );
Parameters vi Instrument handle returned from agb220xa_init( ).
cmd_str Instrument command (cannot exceed 256 bytes in length).
size Length of result in bytes. 2 to 32767.
result[ ] Response from instrument.
This function passes the cmd_str string to the instrument. This entry point passes the string in cmd_str followed by a space and then the integer in value. Note that either an Int16 or 32 can be passed as the Int16 will be promoted.
Syntax ViStatus _VI_FUNC agb220xa_cmdInt(ViSession vi, ViString cmd_str, ViInt32 value);
Parameters vi Instrument handle returned from agb220xa_init( ).
cmd_str Instrument command (cannot exceed 256 bytes in length).
value Parameter for command. -2147483647 to 2147483647.
VXI plug&play Driver · Driver Functions

VXI plug&play Driver Driver Functions
This function passes the cmd_str string to the instrument. This command expects a response that is a definite arbitrary block of 16 bit integers. You specify the cmd_str and size parameters, and get result[ ] and count.
Syntax ViStatus _VI_FUNC agb220xa_cmdInt16Arr_Q(ViSession vi, ViString cmd_str, ViInt32 size, ViInt16_VI_FAR result[ ], ViPInt32 count);
Parameters vi Instrument handle returned from agb220xa_init( ).
cmd_str Instrument command (cannot exceed 256 bytes in length).
size Size of result[ ] (number of items in the array). 1 to 2147483647.
result[ ] Response from instrument.
count Count of valid items in result[ ].
This function passes the cmd_str string to the instrument. This command expects a response that can be returned as a 16 bit integer.
Syntax ViStatus _VI_FUNC agb220xa_cmdInt16_Q(ViSession vi, ViString cmd_str, ViPInt16 result);
Parameters vi Instrument handle returned from agb220xa_init( ).
cmd_str Instrument command (cannot exceed 256 bytes in length).
result Response from instrument.

This function passes the cmd_str string to the instrument. This command expects a response that is a definite arbitrary block of 32 bit integers. You specify the cmd_str and size parameters, and get result[ ] and count.
Syntax ViStatus _VI_FUNC agb220xa_cmdInt32Arr_Q(ViSession vi, ViString cmd_str, ViInt32 size, ViInt32_VI_FAR result[ ], ViPInt32 count);
Parameters vi Instrument handle returned from agb220xa_init( ).
cmd_str Instrument command (cannot exceed 256 bytes in length).
size Size of result[ ] (number of items in the array). 1 to 2147483647.
result[ ] Response from instrument.
count Count of valid items in result[ ].
This function passes the cmd_str string to the instrument. This command expects a response that can be returned as a 32 bit integer.
Syntax ViStatus _VI_FUNC agb220xa_cmdInt32_Q(ViSession vi, ViString cmd_str, ViPInt32 result);
Parameters vi Instrument handle returned from agb220xa_init( ).
cmd_str Instrument command (cannot exceed 256 bytes in length).
result Response from instrument.
VXI plug&play Driver · Driver Functions

VXI plug&play Driver Driver Functions
This function passes the cmd_str string to the instrument. This entry point passes the string in cmd_str followed by a space and then the real in value. Note that either an Real32 or 64 can be passed as the Real32 will be promoted.
Syntax ViStatus _VI_FUNC agb220xa_cmdReal(ViSession vi, ViString cmd_str, ViReal64 value);
Parameters vi Instrument handle returned from agb220xa_init( ).
cmd_str Instrument command (cannot exceed 256 bytes in length).
value Parameter for command. -1E+300 to 1E+300.
This function passes the cmd_str string to the instrument. This command expects a response that is a definite arbitrary block of 64 bit real. You specify the cmd_str and size parameters, and get result[ ] and count.
Syntax ViStatus _VI_FUNC agb220xa_cmdReal64Arr_Q(ViSession vi, ViString cmd_str, ViInt32 size, ViReal64_VI_FAR result[ ], ViPInt32 count);
Parameters vi Instrument handle returned from agb220xa_init( ).
cmd_str Instrument command (cannot exceed 256 bytes in length).
size Size of result[ ] (number of items in the array). 1 to 2147483647.
result[ ] Response from instrument.
count Count of valid items in result[ ].

This function passes the cmd_str string to the instrument. This command expects a response that can be returned as a 64 bit real.
Syntax ViStatus _VI_FUNC agb220xa_cmdReal64_Q(ViSession vi, ViString cmd_str, ViPReal64 result);
Parameters vi Instrument handle returned from agb220xa_init( ).
cmd_str Instrument command (cannot exceed 256 bytes in length).
result Response from instrument.
This function passes the cmd_str string to the instrument. This entry point will wait for a response which must be a string (character data). You specify the cmd_str and size parameters, and get result[ ].
Syntax ViStatus _VI_FUNC agb220xa_cmdString_Q(ViSession vi, ViString cmd_str, ViInt32 size, ViChar_VI_FAR result[ ] );
Parameters vi Instrument handle returned from agb220xa_init( ).
cmd_str Instrument command (cannot exceed 256 bytes in length).
size Length of result in bytes. 2 to 32767.
result[ ] Response from instrument.
VXI plug&play Driver · Driver Functions

VXI plug&play Driver Driver Functions
This function compensates capacitance/conductance data measured by the Agilent 4284A LCR meter, and returns compensation results. Before this function is executed, a compensation data file must be specified by using the agb220xa_selectCompenFile function. The file must contain the appropriate compensation coefficients for your measurement environment. For obtaining the compensation coefficients for your environment and creating the compensation data file, see “Capacitance Compensation” on page 4-20.
Syntax agb220xa_compenC(ViSession vi, ViReal64 frequency, ViReal64 raw_c, ViReal64 raw_g, ViPReal64 compen_c, ViPReal64 compen_g);
Parameters vi Instrument handle returned from agb220xa_init( ).
frequency Measurement frequency (in Hz). 1E3 (1 kHz) to 1E6 (1 MHz).
raw_c Capacitance (in F) measured by the 4284A.
raw_g Conductance (in S) measured by the 4284A.
compen_c Capacitance compensation result (in F). Returned value.
compen_g Conductance compensation result (in S). Returned value.
The function sets connection rule and connection sequence for the specified card.
Syntax ViStatus _VI_FUNC agb220xa_connRuleSeq(ViSession vi, ViInt16 cardno_ruleseq, ViInt16 rule, ViInt16 sequence);
Parameters vi Instrument handle returned from agb220xa_init( ).
cardno_ruleseq Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5 (all card) in the normal configuration mode, or 0 (all card in the auto configuration mode). For the configuration mode, see agb220xa_func.
rule Connection rule. 0 (free route) or 1 (single route).
sequence Connection sequence. 0, 1, or 2. See below.
1 (break before make)
2 (make before break)
0 (no sequence)

This function sets the couple ports which are used for making kelvin connections on the specified card. The specified input port number will be coupled with the next input port and two output ports. For each card, you may setup the same or different couple ports. This command overwrites the previous couple port setting for the card.
The couple mode is controlled by the agb220xa_coupleState function.
If the bias input port and a couple port have been assigned to the same input port, the bias mode and the couple mode cannot be used in parallel.
If the ground input port and a couple port have been assigned to the same input port, the ground mode and the couple mode cannot be used in parallel.
If a ground enabled input port and a couple port have been assigned to the same input port, the ground mode and the couple mode cannot be used in parallel.
Syntax ViStatus _VI_FUNC agb220xa_couplePort(ViSession vi, ViInt16 cport_cardno, ViInt16 port1, ViInt16 port3, ViInt16 port5, ViInt16 port7, ViInt16 port9, ViInt16 port11, ViInt16 port13);
Parameters vi Instrument handle returned from agb220xa_init( ).
cport_cardno Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5 (all card) in the normal configuration mode, or 0 (all card in the auto configuration mode). For the configuration mode, see agb220xa_func.
port1 Couple port by the input ports 1 and 2. 0 (disable) or 1 (enable).
port3 Couple port by the input ports 3 and 4. 0 (disable) or 1 (enable).
port5 Couple port by the input ports 5 and 6. 0 (disable) or 1 (enable).
port7 Couple port by the input ports 7 and 8. 0 (disable) or 1 (enable).
port9 Couple port by the input ports 9 and 10. 0 (disable) or 1 (enable).
port11 Couple port by the input ports 11 and 12. 0 (disable) or 1 (enable).
VXI plug&play Driver · Driver Functions

VXI plug&play Driver Driver Functions
port13 Couple port by the input ports 13 and 14. 0 (disable) or 1 (enable).
This function controls the couple mode for the specified card.
Syntax ViStatus _VI_FUNC agb220xa_coupleState(ViSession vi, ViInt16 cstate_cardno, ViInt16 couple_state);
Parameters vi Instrument handle returned from agb220xa_init( ).
cstate_cardno Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5 (all card) in the normal configuration mode, or 0 (all card in the auto configuration mode). For the configuration mode, see agb220xa_func.
couple_state Couple mode. 0 (OFF) or 1 (ON).
This function sends a device clear (DCL) to the instrument.
A device clear will abort the present operation and enable the instrument to accept a new command or query.
This is particularly useful in situations where it is not possible to determine the instrument state. In this case, it is customary to send a device clear before issuing a new instrument driver function. The device clear ensures that the instrument will be able to begin processing the new commands.
Syntax ViStatus _VI_FUNC agb220xa_dcl(ViSession vi);
Parameters vi Instrument handle returned from agb220xa_init( ).
This function detects the input ports connected to the Kelvin cable, and assigns them as the input couple ports that will be used for the Kelvin connection. The input couple port setting is effective for the all cards. This command overwrites the previous couple port setting.
Syntax ViStatus _VI_FUNC agb220xa_detectCouplePort(ViSession vi);

Parameters vi Instrument handle returned from agb220xa_init( ).
This function translates the error return value from an instrument driver function to a readable string.
Syntax ViStatus _VI_FUNC agb220xa_error_message(ViSession vi, ViStatus error_number, ViChar_VI_FAR message[ ] );
Parameters vi Instrument handle returned from agb220xa_init( ).
error_number Error return value from the driver function.
message[ ] Error message string. This is limited to 256 characters.
This function returns the error numbers and corresponding error messages in the error queue of a instrument. See Chapter 9 for a listing of the instrument error numbers and messages.
Instrument errors may occur when you places the instrument in a bad state such as sending an invalid sequence of coupled commands. Instrument errors can be detected by polling. Automatic polling can be accomplished by using the agb220xa_errorQueryDetect function.
Syntax ViStatus _VI_FUNC agb220xa_error_query(ViSession vi, ViPInt32 error_number, ViChar_VI_FAR error_message[ ] );
Parameters vi Instrument handle returned from agb220xa_init( ).
error_number Instrument’s error code.
error_message[ ] Instrument’s error message. This is limited to 256 characters.
VXI plug&play Driver · Driver Functions

VXI plug&play Driver Driver Functions
This function enables or disables automatic instrument error checking.
If automatic error checking is enabled then the driver will query the instrument for an error at the end of each function call.
Syntax ViStatus _VI_FUNC agb220xa_errorQueryDetect(ViSession vi, ViBoolean errorQueryDetect);
Parameters vi Instrument handle returned from agb220xa_init( ).
errorQueryDetect Error checking enable (VI_TRUE) or disable (VI_FALSE).
This function indicates if automatic instrument error detection is enabled or disabled.
Syntax ViStatus _VI_FUNC agb220xa_errorQueryDetect_Q(ViSession vi, ViPBoolean pErrDetect);
Parameters vi Instrument handle returned from agb220xa_init( ).
pErrDetect Error checking enable (VI_TRUE) or disable (VI_FALSE).

This function returns the contents of the ESR register. The driver returns the equivalent messages (see Parameters).
Syntax ViStatus _VI_FUNC agb220xa_esr_Q(ViSession vi, ViChar_VI_FAR errstr[ ] );
Parameters vi Instrument handle returned from agb220xa_init( ).
errstr[ ] Response from instrument.
2 “ESR_RQL”
4 “ESR_QYE_ERROR”
8 “ESR_DEVICE_DEPENDENT_ERROR”
16 “ESR_EXECUTION_ERROR”
32 “ESR_COMMAND_ERROR”
64 “ESR_URQ”
128 “ESR_PON”
OTHERS “ESR_MULTI_EVENT”
agb220xa_func
This function is used to set the channel configuration to the auto configuration mode or the normal configuration mode.
Syntax ViStatus _VI_FUNC agb220xa_func(ViSession vi, ViInt16 channel_config);
Parameters vi Instrument handle returned from agb220xa_init( ).
channel_config Configuration mode. 0 (auto) or 1 (normal).
VXI plug&play Driver · Driver Functions · Bit Value Message

VXI plug&play Driver Driver Functions
This function ground-enables or disables all the output ports of the specified card.
Syntax ViStatus _VI_FUNC agb220xa_groundChanCard(ViSession vi, ViInt16 disable_enable, ViInt16 gnd_cardno);
Parameters vi Instrument handle returned from agb220xa_init( ).
disable_enable Status of the card, ground enabled or ground disabled.
gnd_cardno Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5 (all card) in the normal configuration mode, or 0 (all card in the auto configuration mode). For the configuration mode, see agb220xa_func.
This function ground-enables or disables all the output ports specified by the gndchan_list.
The parameter “gndchan_list” is an array of integers with each integer representing one channel. The last number of the “gndchan_list” should be “0” (numeric zero) to identify the end of the list. The maximum number of channels that can be specified by the list is 100.
Syntax ViStatus _VI_FUNC agb220xa_groundChanList(ViSession vi, ViInt16 gndchan_disen, ViInt32_VI_FAR gndchan_list[ ] );
Parameters vi Instrument handle returned from agb220xa_init( ).
gndchan_disen Status of the port, ground enabled or ground disabled.
gndchan_list[ ] Channel numbers. 5 digits integer. ABCDE. where A: card number, BC: input port number, DE: output port number. Top zero(s) can be ignored. For example, if A=0, BC=01, and DE=01, channel number should be 101 instead of 00101.

This function will query the instrument for the status, ground enabled or disabled, for the channels given in the list.
The parameter “gndchan_list” is an array of integers with each integer representing one channel. The last number of the “gndchan_list” should be “0” (numeric zero) to identify the end of the list. The maximum number of channels that can be specified by the list is 100.
The “gnd_status” parameter is an array of integers containing the return values of the query. The “gnd_status” array returned will correspond one to one with “gndchan_list” parameter.
Syntax ViStatus _VI_FUNC agb220xa_groundChanList_Q(ViSession vi, ViInt16 gnd_disen, ViInt32_VI_FAR gndchan_list[ ], ViInt32_VI_FAR gnd_status[ ] );
Parameters vi Instrument handle returned from agb220xa_init( ).
gnd_disen Query type.
gndchan_list[ ] Channel numbers to check the status. 5 digits integer. ABCDE. where A: card number, BC: input port number, DE: output port number. Top zero(s) can be ignored. For example, if A=0, BC=01, and DE=01, channel number should be 101 instead of 00101.
gnd_status[ ] Status of the channels given in the gndchan_list. Returned value depends on the setting of gnd_disen as shown below:
when gnd_disen=0, 0 means ground disabled, 1 means enabled.
when gnd_disen=1, 0 means ground enabled, 1 means disabled.
VXI plug&play Driver · Driver Functions

VXI plug&play Driver Driver Functions
This function will select which input port is the ground port on the specified card. For each card, you can specify the same or different ground port.
Syntax ViStatus _VI_FUNC agb220xa_groundPort(ViSession vi, ViInt16 gport_cardno, ViInt16 gnd_port);
Parameters vi Instrument handle returned from agb220xa_init( ).
gport_cardno Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5 (all card) in the normal configuration mode, or 0 (all card in the auto configuration mode). For the configuration mode, see agb220xa_func.
gnd_port Input port number to be set to the ground port. 1 to 14 (input port 1 to input port 14).
The input ground port and a ground enabled input port cannot be assigned to the same input port.
If the ground input port and a couple port have been assigned to the same input port, the ground mode and the couple mode cannot be used in parallel.
This function controls the ground mode for the specified card. When ground mode is ON, the input ground port is connected to all ground enabled input ports and output ports that are not connected to any other input ports. Ground disabled output ports are never connected to the input ground port when ground mode is ON.
The ground mode cannot be set to ON when the bias mode is ON.
Syntax ViStatus _VI_FUNC agb220xa_groundState(ViSession vi, ViInt16 gstate_cardno, ViInt16 state);
Parameters vi Instrument handle returned from agb220xa_init( ).
gstate_cardno Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5 (all card) in the normal configuration mode, or 0 (all card in the auto configuration mode). For the configuration mode, see agb220xa_func.
state Ground mode. 0 (OFF) or 1 (ON).

This function initializes the software connection to the instrument and optionally verifies that instrument is in the system. In addition, it may perform any necessary actions to place the instrument in its reset state.
If the agb220xa_init function encounters an error, then the value of the vi output parameter will be VI_NULL.
Syntax ViStatus _VI_FUNC agb220xa_init(ViRsrc InstrDesc, ViBoolean id_query, ViBoolean do_reset, ViPSession vi);
Parameters InstrDesc Instrument description. Examples; GPIB0::1::INSTR.
id_query VI_TRUE (to perform In-System Verification), or VI_FALSE (do not perform In-System Verification).
do_reset VI_TRUE (to perform reset operation), or VI_FALSE (do not perform reset operation).
vi Instrument handle. This is VI_NULL if an error occurred during the initialization.
This function does the *OPC? common command.
Syntax ViStatus _VI_FUNC agb220xa_opc_Q(ViSession vi, ViPBoolean result);
Parameters vi Instrument handle returned from agb220xa_init( ).
result VI_TRUE (Operation complete), or VI_FALSE (Operation is pending).
VXI plug&play Driver · Driver Functions

VXI plug&play Driver Driver Functions
This function will disconnect all input ports from all output ports for the specified card.
If bias mode is ON, the input bias port is connected to all bias enabled output ports.
If ground mode is ON, the input ground port is connected to all ground enabled input ports and output ports.
Syntax ViStatus _VI_FUNC agb220xa_openCard(ViSession vi, ViInt16 open_cardno);
Parameters vi Instrument handle returned from agb220xa_init( ).
open_cardno Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5 (all card) in the normal configuration mode, or 0 (all card in the auto configuration mode). For the configuration mode, see agb220xa_func.
This function will disconnect the input ports from the output ports specified by the channel list.
The parameter “openchan_list” is an array of integers with each integer representing one channel. The last number of the “openchan_list” should be “0” (numeric zero) to identify the end of the list. The maximum number of channels that can be specified by the list is 100.
Syntax ViStatus _VI_FUNC agb220xa_openList(ViSession vi, ViInt32_VI_FAR openchan_list[ ] );
Parameters vi Instrument handle returned from agb220xa_init( ).
openchan_list[ ] Channel numbers to disconnect. 5 digits integer. ABCDE. where A: card number, BC: input port number, DE: output port number. Top zero(s) can be ignored. For example, if A=0, BC=01, and DE=01, channel number should be 101 instead of 00101.

This function will query the instrument for the channels open given in the “openchan_list”.
The parameter “openchan_list” is an array of integers with each integer representing one channel. The last number of the “openchan_list” should be “0” (numeric zero) to identify the end of the list. The maximum number of channels that can be specified by the list is 100.
The “open_status” parameter is an array of integers containing the return values of the query. The “open_status” array returned will correspond one to one with “openchan_list” parameter.
Syntax ViStatus _VI_FUNC agb220xa_openList_Q(ViSession vi, ViInt32_VI_FAR openchan_list[ ], ViInt32_VI_FAR open_status[ ] );
Parameters vi Instrument handle returned from agb220xa_init( ).
openchan_list[ ] Channel numbers to know the open status. 5 digits integer. ABCDE. where A: card number, BC: input port number, DE: output port number. Top zero(s) can be ignored. For example, if A=0, BC=01, and DE=01, channel number should be 101 instead of 00101.
open_status[ ] Status of the channels given in the openchan_list. 1 (opened) or 0 (closed).
This function returns the contents of the status byte register.
Syntax ViStatus _VI_FUNC agb220xa_readStatusByte_Q(ViSession vi, ViPInt16 statusByte);
Parameters vi Instrument handle returned from agb220xa_init( ).
statusByte The contents of the status byte are returned in this parameter.
VXI plug&play Driver · Driver Functions

VXI plug&play Driver Driver Functions
This function places the instrument in a default state. Before issuing this function, it may be necessary to send a device clear to ensure that the instrument can execute a reset. A device clear can be issued by invoking agb220xa_dcl function.
Syntax ViStatus _VI_FUNC agb220xa_reset(ViSession vi);
Parameters vi Instrument handle returned from agb220xa_init( ).
This function returns the driver revision and the instrument firmware revision.
Syntax ViStatus _VI_FUNC agb220xa_revision_query(ViSession vi, ViChar_VI_FAR driver_rev[ ] , ViChar_VI_FAR instr_rev[ ] );
Parameters vi Instrument handle returned from agb220xa_init( ).
driver_rev[ ] Instrument driver revision. This is limited to 256 characters.
instr_rev[ ] Instrument firmware revision. This is limited to 256 characters.
This function specifies the compensation data file used by the agb220xa_compenC function. The file must contain the appropriate compensation coefficients for your measurement environment.
For obtaining the compensation coefficients for your environment and creating the compensation data file, see “Capacitance Compensation” on page 4-20.
Syntax ViStatus _VI_FUNC agb220xa_selectCompenFile(ViSession vi, ViString file_name);
Parameters vi Instrument handle returned from agb220xa_init( ).
file_name Compensation data file name. Use absolute path. If the value is NULL string, the default data is used.

This function causes the instrument to perform a self-test and returns the result of that self-test. This is used to verify that an instrument is operating properly. A failure may indicate a potential hardware problem.
Syntax ViStatus _VI_FUNC agb220xa_self_test(ViSession vi, ViPInt16 test_result, ViChar_VI_FAR test_message[ ] );
Parameters vi Instrument handle returned from agb220xa_init( ).
test_result Numeric result from self-test operation. 0: No error.
test_message[ ] Self-test status message. This is limited to 256 characters.
This function clears the test result of the specified diagnostics item.
Syntax ViStatus _VI_FUNC agb220xa_testClear(ViSession vi, ViInt16 framecard_clear);
Parameters vi Instrument handle returned from agb220xa_init( ).
framecard_clear Test result to clear. 0 to 10. See below.
VXI plug&play Driver · Driver Functions

VXI plug&play Driver Driver Functions
This function performs the diagnostics and returns the test result.
For details of each test, see “Selftest Menu” on page 3-34.
Syntax ViStatus _VI_FUNC agb220xa_testExec_Q(ViSession vi, ViInt16 framecard_exec, ViPInt16 exec_result);
Parameters vi Instrument handle returned from agb220xa_init( ).
framecard_exec Test item to perform. 1 to 10. See below.
exec_result Test result. 0: pass, 1: fail.
This function sets a minimum timeout value for driver I/O transactions in milliseconds. The default timeout period is 2 seconds.
Syntax ViStatus_VI_FUNC agb220xa_timeOut(ViSession vi, ViInt32 timeOut);
Parameters vi Instrument handle returned from agb220xa_init( ).
timeOut I/O timeout value for all functions in the driver. in milliseconds. 0 to 2147483647.

This function returns the timeout value for driver I/O transactions in milliseconds.
Syntax ViStatus_VI_FUNC agb220xa_timeOut_Q(ViSession vi, ViPInt32 pTimeOut);
Parameters vi Instrument handle returned from agb220xa_init( ).
pTimeOut Minimum timeout period that the driver can be set to, in milliseconds.
This function ground-enables the specified input ports of the specified card.
The parameter “unused_port” is an array of integers with each integer representing one port. The last number of the “unused_port” should be “0” (numeric zero) to identify the end of the array. The maximum number of ports that can be specified by the array is agb220xa_UNUSEDPORT_MAX.
Syntax ViStatus _VI_FUNC agb220xa_unusedPort(ViSession vi, ViInt16 unused_cardno, ViInt32_VI_FAR unused_port[ ] );
Parameters vi Instrument handle returned from agb220xa_init( ).
unused_cardno Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5 (all card) in the normal configuration mode, or 0 (all card in the auto configuration mode). For the configuration mode, see agb220xa_func.
unused_port[ ] Input ports to be ground enabled. 1 to 9 are available. Multiple port numbers can be set.
The input ground port and a ground enabled input port cannot be assigned to the same input port.
If the ground enabled input port and a couple port have been assigned to the same input port, the ground mode and the couple mode cannot be used in parallel.
VXI plug&play Driver · Driver Functions

VXI plug&play Driver Driver Functions


This chapter lists and describes the error messages for Agilent B2200. An error message consists of an error number and message.
There are two types of error messages:
Negative error numbers (Command Error, Execution Error, Device-Dependent Error, and Query Error) are standard SCPI errors.
Positive error numbers are the B2200 specific errors.
Error messages are classified by error number as listed in the following table:
Error Range Error Category
-100 to -199 Command Error
-200 to -299 Execution Error
-300 to -399 Device-Dependent Error
-400 to -499 Query Error
100 to 32767 B2200 Specific Error
When an error occurs, the corresponding bit is set in the Standard Event Status Register (see Chapter 5):
Error Category Standard Event Status Register Bit
B2200 Specific Error bit3
Also, the error number and message are placed in the error queue, which can be read by the :SYSTem:ERRor? query command. The error queue is cleared by the common command *CLS, and when power is turned on. For these commands, see Chapter 5.

Error Messages Standard SCPI Error Messages
Standard SCPI error messages have negative error numbers. The following are the standard SCPI error messages for the B2200:
If syntax of SCPI command is not valid, a -1XX error occurs. The error number and message are placed in the error queue, and bit5 of the Standard Event Status Register is set.
A SCPI command consists of a command header and zero or more parameters. The following are example SCPI commands:
:ROUT:OPEN:CARD is the command header, and 1 is the parameter.
A subsystem command header consists of mnemonics (keywords) separated by colons. Query commands have a question mark (?) at end of last mnemonic.
*SRE is the command header, and 48 is the parameter.
A common command header consists of an asterisk (*) followed by 3 characters. Query commands have a question mark (?) at end of header.
Error Number Error Message and Description
-100 Command error
Generic syntax error that cannot be determined more specifically.
-101 Invalid character
A syntax element contains a character that is invalid for that type of element; for example, a header containing an ampersand.
-102 Syntax error
An unrecognized command or data type was received; for example, a string was received when the B2200 does not accept strings.

Error Messages Standard SCPI Error Messages
Error Number Error Message and Description
-103 Invalid separator
An illegal character was received when a separator was expected; for example, the semicolon was omitted between multiple commands in a program message.
-104 Data type error
An improper data type was received; for example, numeric data was expected but string data was received.
-105 GET not allowed
A group execute trigger was received within a program message.
-108 Parameter not allowed
Too many parameters for the command were received.
-109 Missing parameter
Fewer parameters were received than required for the command.
-110 Command header error
An error was detected in the header. This error message is reported if the B2200 cannot determine the more specific header errors -111 through -114.
-111 Header separator error
An illegal character for a header separator was received; for example, no white space between the command header and parameter.
-112 Program mnemonic too long
A keyword in the command header contains more than twelve characters.
-113 Undefined header
An undefined command header was received; for example, *XYZ or :ROUT:OPEN:COOD.

Error Messages Standard SCPI Error Messages
Error Number Error Message and Description
-120 Numeric data error
An error was detected in a numeric parameter (including the non-decimal numeric types). This error message is reported when the B2200 cannot determine the more specific errors -121 through -128.
-121 Invalid character in number
An invalid character for the parameter was received; for example, an alphacharacter was received when the parameter type was decimal numeric.
-123 Exponent too large
The magnitude of the exponent for a numeric parameter was larger than 32000.
-124 Too many digits
The mantissa of a decimal numeric parameter contained more than 255 digits excluding leading zeros.
-128 Numeric data not allowed
Numeric data is not allowed in this position for this command.
-138 Suffix not allowed
A suffix was received after a numeric parameter. For the B2200, no parameters have suffix.
-140 Character data error
An error was detected in a character parameter. This error message is reported if the B2200 cannot determine the more specific errors -141 through -148.
-141 Invalid character data
Either the character parameter contains an invalid character or the particular element received is not valid for the command.
-144 Character data too long
The character parameter contains more than 12 characters.

Error Messages Standard SCPI Error Messages
Error Number Error Message and Description
-148 Character data not allowed
A character parameter is not allowed for this position.
-150 String data error
An error was detected in a string parameter. This error is reported if the B2200 cannot determine a more specific error -151 and -158.
-151 Invalid string data
An invalid string parameter data was received; for example, an END message was received before the terminal quote character.
-158 String data not allowed
A string parameter data was received but was not allowed at this point.
-160 Block data error
An error was detected in a block data. This error is reported if the B2200 cannot determine more specific errors -161 and -168.
-161 Invalid block data
An invalid block data was received; for example, an END message was received before the length was satisfied.
-168 Block data not allowed
A legal block data was received but was not allowed at this point.
-170 Expression error
An error was detected in an expression. This error is reported if the B2200 cannot determine more specific errors -171 and -178.
-171 Invalid expression
The expression was invalid; for example, unmatched parentheses or an illegal character.
-178 Expression data not allowed
An expression was received but was not allowed at this point.

Error Messages Standard SCPI Error Messages
If syntax of a SCPI command header and parameter is valid, but the command cannot be executed due to some condition of the B2200, a -2XX error occurs. The error number and message are placed in the error queue, and bit4 of the Standard Event Status Register is set.
Error Number Error Message and Description
-200 Execution error
Generic execution error that cannot be determined more specifically.
-220 Parameter error
The parameter value is valid, but not executable due to some condition of the B2200. This error occurs if more specific errors -221 through -224 cannot be determined.
-222 Data out of range
A valid parameter setting could not be executed because interpreted value was out of range as defined by the B2200.
-223 Too much data
A valid parameter setting (block, expression, or string type) could not be executed because parameter contained more data than the B2200 could handle due to insufficient memory or other requirements.
-224 Illegal parameter value
A valid parameter setting (where exact value from list of possibles was expected) could not be executed due to present B2200 state.
-260 Expression error
An expression related error occurred.

Error Messages Standard SCPI Error Messages
-3XX errors indicate that an B2200 operation did not properly complete, possibly due to an abnormal hardware or firmware condition. These negative codes are SCPI defined. For the device-dependent positive error codes, see “B2200 Specific Error Messages” on page 7-10. The positive codes are not SCPI defined.
For these errors, an error number and message are placed in the error queue, and bit3 of the Standard Event Status Register is set.
Error Number Error Message and Description
-300 Device-specific error
Generic device-dependent error for the B2200 that cannot be determined more specifically.
-311 Memory error
An error was detected in the B2200's memory.
-350 Queue overflow
If error queue is full, this code is entered into the error queue instead of the code that caused the error. This code indicates that there was no room in the error queue, so an error occurred but was not recorded.

Error Messages Standard SCPI Error Messages
If the output queue control of the B2200 detects one of following problems, a -4XX error occurs:.
If this type of error occurs, the error number and message are placed in the error queue, and bit2 of the Standard Event Status Register is set.
Error Number Error Message and Description
-400 Query error
Generic query error for the B2200 that cannot be determined more specifically.
-410 Query INTERRUPTED
A condition causing an INTERRUPTED query error occurred; for example, a query followed by DAB or GET before a response was completely sent.
-420 Query UNTERMINATED
A condition causing an UNTERMINATED query error occurred; for example, the B2200 was addressed to talk and an incomplete program message was received.
-430 Query DEADLOCKED
A condition causing a DEADLOCKED query error occurred; for example, both input buffer and output buffer are full and the B2200 cannot continue.
-440 Query UNTERMINATED after indefinite response
A query was received in the same program message after a query requesting an indefinite length response was executed.

Error Messages B2200 Specific Error Messages
These are the B2200-specific errors that are not defined by SCPI. These errors indicate that an B2200 operation did not properly complete due to card, channel, port, or mode errors.
For the SCPI defined device-dependent codes, see “Device-Dependent Errors” on page 7-8.
For these errors, an error number and message are placed in the error queue, and bit3 of the Standard Event Status Register is set.
The B2200 specific error messages have positive error numbers.

Error Messages B2200 Specific Error Messages
Error Number Error Message and Description
2000 Invalid card number
Wrong card number is specified in card number or channel list parameter. Make sure card is properly installed in the correct B2200 slot.
2001 Invalid channel number
Wrong channel number is specified in the channel list. Confirm the channel number, the card configuration, and the configuration mode of the B2200.
2006 Command not supported on this card
Command was used that is not supported by the card.
2009 Too many channels in channel list
Too many channels are specified in the channel list (maximum 120).
2011 Empty channel list
No channel list is specified.
2012 Invalid channel range
Wrong channel list is defined. Confirm the syntax of the channel list, the card configuration, and the configuration mode of the B2200.
2002 Unsupported Module · 2003 Unsupported Configuration

Error Messages B2200 Specific Error Messages
Error Number Error Message and Description
3000 Card0 initialization fail
The B2200 may be defective. Contact your nearest Agilent Technologies service center.
3001 Card1 initialization fail
The B2200 or card installed in slot 1 of the B2200 may be defective.
3002 Card2 initialization fail
The B2200 or card installed in slot 2 of the B2200 may be defective.
3003 Card3 initialization fail
The B2200 or card installed in slot 3 of the B2200 may be defective.
3004 Card4 initialization fail
The B2200 or card installed in slot 4 of the B2200 may be defective.
3011 Bad couple port number
For the Input Port No. for the couple port, only 1, 3, 5, 7, or 9 are allowed.
3012 Bad bias port number
For Input Port No. for Bias Port, only 1 to 10 are allowed.
3013 Cannot connect multiple channels in SROUTe mode
For the single connection rule, an input port can be connected to only one output port, and an output port can be connected to only one input port.
3014 Cannot directly specify Bias Port channel
Relays on the input bias port cannot be controlled directly when the bias mode is ON.

Error Messages B2200 Specific Error Messages
Error Number Error Message and Description
3017 Too many relays closed. Max 52 relays/card.
Too many relays have been closed. To close new relays, open some relays. Maximum 52 relays can be closed for each module.
3018 Can't change to ACONfig mode. Check card config.
The present card configuration of the B2200 does not allow the auto configuration mode. Change the module configuration.
3019 Cannot use same port for Couple and Bias
The couple mode and the bias mode cannot be used in parallel when a couple port and the input bias port have been assigned to the same input port.
3020 Bad auto ground port number
Specify the output ports effective for the input ground port. The port number must be 1 to 14
3021 Bad unused port number
Specify the input ports effective for the ground enabled input port (unused port). The port number must be 1 to 9.
3022 Cannot directly specify auto ground port channel
Relays on the input ground port cannot be controlled directly when the ground mode is ON.
3023 Cannot directly specify unused port channel
Relays on the ground enabled input port (unused port) cannot be controlled directly when the ground mode is ON.
3024 Cannot use same port for Couple and Auto Ground
The couple mode and the ground mode cannot be used in parallel when a couple port and the input ground port have been assigned to the same input port.
3025 Cannot use same port for Unused and Auto Ground
The input ground port and the ground enabled input port (unused port) cannot be assigned to the same input port.

Error Messages B2200 Specific Error Messages
Error Number Error Message and Description
3026 Cannot use same port for Unused and Couple
The couple mode and the ground mode cannot be used in parallel when a couple port and the ground enabled input port (unused port) have been assigned to the same input port.
3027 Cannot use Unused Port during Auto Ground Mode ON
The ground enabled input port (unused port) must be opened during the ground mode is ON.
3030 Bad input port number
The input port number must be 1 to 14.
3031 Bad output port number
The output port number must be 1 to 12, 24, 36, or 48. Maximum number depends on the switch module configuration.
3032 Bad setting memory number
The internal memory number for the setup data must be 1 to 8.
3033 EEPROM programming failure
Cannot change the GPIB address or update internal memory data. Contact your nearest Agilent Technologies service center.
3034 EEPROM reading failure
Cannot read the GPIB address or internal memory data. Contact your nearest Agilent Technologies service center.
3035 Setting memory data is invalid
Cannot read the setup data in the internal memory. The data has been broken. Delete the data.
3036 Cannot load this setting data in this configuration
Cannot read the setup data in the internal memory due to the configuration mismatch. Delete the data or leave it until restoring the module configuration.
What the fault codes shown on the display mean and how to clear them.
| Code | Meaning | Fix | |
|---|---|---|---|
| -300 | Device-specific error Generic device-dependent error for the B2200 that cannot be determined more specifically. | p.228 | |
| -311 | Memory error An error was detected in the B2200's memory. | p.228 | |
| -350 | Queue overflow If error queue is full, this code is entered into the error queue instead of the code that caused the error. This code indicates that there was no room in the error queue, so an error occurred but was not recorded. | p.228 | |
| -400 | Query error Generic query error for the B2200 that cannot be determined more specifically. | p.229 | |
| -410 | Query INTERRUPTED A condition causing an INTERRUPTED query error occurred; for example, a query followed by DAB or GET before a response was completely sent. | p.229 | |
| -420 | Query UNTERMINATED A condition causing an UNTERMINATED query error occurred; for example, the B2200 was addressed to talk and an incomplete program message was received. | p.229 | |
| -430 | Query DEADLOCKED A condition causing a DEADLOCKED query error occurred; for example, both input buffer and output buffer are full and the B2200 cannot continue. | p.229 | |
| -440 | Query UNTERMINATED after indefinite response A query was received in the same program message after a query requesting an indefinite length response was executed. | p.229 |