Agilent Technologies Part Number 89441-90076
For instruments with firmware version A.08.00 Printed in U.S.A.
Print Date: June 2000 © Agilent Technologies 1994, 1995, 2000. All rights reserved. 8600 Soper Hill Road Everett, Washington 98205-1298 U.S.A.
This software and documentation is based in part on the Fourth Berkeley Software Distribution under license from The Regents of the University of California. We acknowledge the following individuals and institutions for their role in the development: The Regents of the University of California .



Front Panel 1-A softkey’s function changes as different menus are displayed. Its current function is determined by the video label to its left, on the analyzer’s screen.
11-Use the SYSTEM hardkeys and their menus to control various system functions (online help, plotting, presetting, and so on).
12-Use the MEASUREMENT hardkeys and their menus to control the analyzer’s receiver and source, and to specify other measurement parameters.
2-The analyzer’s screen is divided into two main areas. The menu area, a narrow column at the screen’s right edge, displays softkey labels. The data area, the remaining portion of the screen, displays traces and other data.
13-The REMOTE OPERATION hardkey and LED indicators allow you to set up and monitor the activity of remote devices.
3-The POWER switch turns the analyzer on and off.
14-Use the MARKER hardkeys and their menus to control marker positioning and marker functions.
4-Use a 3.5 inch flexible disk (DS,HD) in this disk drive to save your work.
5-The KEYBOARD connector allows you to attach an optional keyboard to the analyzer. The keyboard is most useful for writing and editing Agilent Instrument BASIC programs.
15-The knob’s primary purpose is to move a marker along the trace. But you can also use it to change values during numeric entry, move a cursor during text entry, or select a hypertext link in help topics
16-Use the Marker/Entry key to determine the knob’s function. With the Marker indicator illuminated the knob moves a marker along the trace. With the Entry indicator illuminated the knob changes numeric entry values.
17-Use the ENTRY hardkeys to change the value of numeric parameters or to enter numeric characters in text strings.
7-The EXT TRIGGER connector lets you provide an external trigger for the analyzer.
8-The PROBE POWER connectors provides power for various Agilent active probes.
18-The optional CHANNEL 2 input connector routes your test signal or DUT output to the analyzer’s receiver. Input impedance is selectable: 50 ohms, 75 ohms, or 1 megohm. For ease of upgrading, the CHANNEL 2 BNC connector is installed even if option AY7 (second input channel) is not installed.
9-The INPUT connector routes your test signal or DUT output to the analyzer’s receiver. Input impedance is selectable: 50 ohms or 75 ohms with option 1D7 (minimum loss pads).
10-Use the DISPLAY hardkeys and their menus to select and manipulate trace data and to select display options for that data.
For more details on the front panel, display the online help topic “Front Panel”. See the chapter “Using Online Help” if you are not familiar with using the online help index.

This page left intentionally blank.

The following general safety precautions must be observed during all phases of operation of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies, Inc. assumes no liability for the customer’s failure to comply with these requirements.
GENERAL This product is a Safety Class 1 instrument (provided with a protective earth terminal). The protective features of this product may be impaired if it is used in a manner not specified in the operation instructions.
All Light Emitting Diodes (LEDs) used in this product are Class 1 LEDs as per IEC 60825-1.
ENVIRONMENTAL CONDITIONS This instrument is intended for indoor use in an installation category II, pollution degree 2 environment. It is designed to operate at a maximum relative humidity of 95% and at altitudes of up to 2000 meters. Refer to the specifications tables for the ac mains voltage requirements and ambient operating temperature range.
BEFORE APPLYING POWER Verify that the product is set to match the available line voltage, the correct fuse is installed, and all safety precautions are taken. Note the instrument’s external markings described under Safety Symbols.
GROUND THE INSTRUMENT To minimize shock hazard, the instrument chassis and cover must be connected to an electrical protective earth ground. The instrument must be connected to the ac power mains through a grounded power cable, with the ground wire firmly connected to an electrical ground (safety ground) at the power outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury.

FUSES Only fuses with the required rated current, voltage, and specified type (normal blow, time delay, etc.) should be used. Do not use repaired fuses or short-circuited fuse holders. To do so could cause a shock or fire hazard.
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE Do not operate the instrument in the presence of flammable gases or fumes.
DO NOT REMOVE THE INSTRUMENT COVER Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be made only by qualified service personnel.
Instruments that appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel. WARNING The WARNING sign denotes a hazard. It calls attention to a procedure, practice, or the like, which, if not correctly performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING sign until the indicated conditions are fully understood and met.

Protective earth (ground) terminal
Frame or chassis terminal
Terminal is at earth potential.
Standby (supply). Units with this symbol are not completely disconnected from ac mains when this switch is off
Safety Symbols · Earth (ground) terminal

Hardkeys Hardkeys are front-panel buttons whose functions are always the same. Hardkeys have a label printed directly on the key. In this book, they are printed like this: [Hardkey].
Softkeys Softkeys are keys whose functions change with the analyzer’s current menu selection. A softkey’s function is indicated by a video label to the left of the key (at the edge of the analyzer’s screen). In this book, softkeys are printed like this: [softkey].
Toggle Softkeys Some softkeys toggle through multiple settings for a parameter. Toggle softkeys have a word highlighted (of a different color) in their label. Repeated presses of a toggle softkey changes which word is highlighted with each press of the softkey. In this book, toggle softkey presses are shown with the requested toggle state in bold type as follows: “Press [key name on]” means “press the softkey [key name] until the selection on is active.”
Shift Functions In addition to their normal labels, keys with blue lettering also have a shift function. This is similar to shift keys on an pocket calculator or the shift function on a typewriter or computer keyboard. Using a shift function is a two-step process. First, press the blue [Shift] key (at this point, the message “shift” appears on the display). Then press the key with the shift function you want to enable. Shift function are printed as two key presses, like this: [Shift] [Shift Function]
Numeric Entries Numeric values may be entered by using the numeric keys in the lower right hand ENTRY area of the analyzer front panel. In this book values which are to be entered from these keys are indicted only as numerals in the text, like this: Press 50, [enter]
Ghosted Softkeys A softkey label may be shown in the menu when it is inactive. This occurs when a softkey function is not appropriate for a particular measurement or not available with the current analyzer configuration. To show that a softkey function is not available, the analyzer ‘’ghosts’’ the inactive softkey label. A ghosted softkey appears less bright than a normal softkey. Settings/values may be changed while they are inactive. If this occurs, the new settings are effective when the configuration changes such that the softkey function becomes active.

This book, “Agilent Technologies 89441A Getting Started Guide”, is designed to help you become comfortable with the Agilent 89441A Vector Signal Analyzers. It provides step-by step examples of how to use this analyzer to perform tasks which you have probably performed with other analyzers. By performing these tasks you will become familiar with many of the basic features—and how those features fit together to perform actual measurements.
This book also contains a chapter to help you prepare the analyzer for use, including instructions for inspecting and installing the analyzer.
To Learn More About the Analyzer You may need to use other books in the analyzer’s manual set. See the “Documentation Roadmap” at the end of this book to learn what each book contains.

This page left intentionally blank


You can learn about your analyzer from online help which is built right into the instrument and is available to you any time you use the analyzer. This section shows you how to use online help to learn about specific keys or topics. You can use online help in conjunction with other documentation to learn about your analyzer in depth, or you can refresh your memory for keys you seldom use. You can use online help while working with your analyzer since online help does not alter the analyzer setup.

1 Enter the online help system: Press [Help]. 2 Display online help for the [Help] hardkey: Press [5] on the numeric keypad. 3 Use the knob or the up-arrow or down-arrow keys to move through the pages. 4 Quit online help: Press [Help]. or Press [0] on the keypad. Take a few moments to read the help overview. It’s only five pages long, and it includes descriptions of advanced features like the index and cross-reference “links” that can help you locate the information you need more quickly.
When you enter the help system it displays help on the last key you pressed. If you have just turned on the analyzer online help for the [Help] key is displayed.
When you quit help, the analyzer restores the display and menu that was displayed before you enabled help. Using online help does not alter your measurement setup.

This legend shows which numeric keys access online help features
Using Online Help

This example displays topics related to triggering. 1 Enter the online help system: Press [Help]. 2 Display help for a hardkey: Press [Trigger]. 3 Use the knob or the up and down arrow keys to page through the topic. 4 Select a softkey topic: Press [trigger type], [IF channel 1]. 5 Quit online help: Press [Help] or Press [0] on the keypad. Pressing [Preset] always returns the analyzer to its preset state. If you press any other key when help is enabled, the analyzer displays a help topic describing the key’s function. For help on the preset state, select “Preset hardkey” from the help index (you will learn how to do this later in this section) or press [Preset] then [Help].

These lines show the name of the selected softkey and the path to its hardkey
Using Online Help

This example displays topics related to saving and recalling. 1 Enter the online help system: Press [Help]. 2 Display help for a hardkey: Press [Save/Recall]. 3 Scroll with the knob to highlight the Math topic. 4 Select that topic: Press [4]. 5 Return to previous topics: Press [7]. 6 Quit online help: Press [Help]. On a given screen full of online help text, there may be several special words (or phrases) that are linked to related topics. Most of these words are underlined to identify them as links, but one is highlighted to identify it as the currently-selected link. The knob allows you to select a different link by moving the highlighting from one link to the next. Once you’ve selected the link you want, press [4] on the keypad to display the related topic.
You can follow links through as many as 20 topics and still return to the original topic. Just press [7] one time for each link you followed, and you’ll return to the original topic via all of the related topics you displayed.

The highlighted link shows what topic is displayed if you press 4
Underlined links show other topics available from this online help topic
Using Online Help

1 Enter the online help system: Press [Help]. 2 Display the index: Press [1]. 3 Turn the knob to select the topic you want help on or for faster paging press and hold the up-arrow or down-arrow keys then use the knob to select a topic. 4 Display the topic: Press [4]. 5 Quit online help: Press [Help]. or Press [0]. The help index contains an alphabetical listing of all help topics. Most topics listed in the index describe the hardkeys and softkeys, but some are of a more general nature. These more general topics are only available via the index or via “links” from related topics. An example appears below–the “Front Panel Tour” topic is only available through the index or the “links”, not by pressing any hardkey or softkey.

You can select any topic in the index by scrolling to highlight it then pressing 4
Using Online Help



This chapter shows you how to make typical noise measurements. In this example, we will be making random noise, band power noise, and signal to noise measurements.

Making Simple Noise Measurements
1 Initialize the analyzer: Press [Preset]. 2 Select a power spectral density measurement: Press [Measurement Data], [PSD] (select ch1 with a 2-channel analyzer). 3 Turn on averaging: Press [Average], [average on]. 4 Start an averaged measurement: Press [Meas Restart]. 5 Use the knob to move the marker along the trace. The display should be similar to the one shown below.
To learn more about the choices you make in this measurement, display online help for the various keys used (see “Using Online Help” if you are not familiar with how to do this).

In this example you are measuring the noise-power of the analyzer’s noise floor. The displayed marker value reflects noise-power normalized to a 1-Hz bandwidth.
Normalized noise measurement

Making Simple Noise Measurements
1 Initialize the analyzer: Press [Preset]. 2 Turn on averaging: Press [Average], [average on]. 3 Start an averaged measurement: Press [Meas Restart]. 4 Turn on the band power markers: Press [Marker Function], [band power markers], [band pwr mkr on], [band power] Press [ResBW/Window], [detector], [sample] Press [Marker Function], [band power markers]. 5 Change the width of the band: Press [band right], [Marker|Entry], then use the knob to move the marker to the desired location. Press [band left], then use the knob to move the marker to the desired location. The display should similar to the one below. The grid lines have been turned off to highlight the band power markers.

In this example you are measuring the power of the analyzer’s noise floor within a defined band. The value displayed in the lower left corner of the display reflects the total power within the frequency band encompassed by the markers. The grid lines have been turned off to highlight the band power markers.
Band power markers · Band power magnitude

Making Simple Noise Measurements
1 Select the baseband receiver mode and initialize the analyzer: Press [Instrument Mode] [receiver] [RF section (0-10 MHz)]. Press [Preset]. 2 Supply a signal from the internal source: Connect the SOURCE output to the INPUT with a BNC cable. Press [Source], [source on], [sine freq], 5, [MHz] 3 Place the marker on the signal peak: Press [Marker⇒], [marker to peak] or Press [Shift], [Marker] 4 Select video averaging: Press [Average], [average on] 5 Turn on the carrier-to-noise marker: Press [Marker Function], [band power markers], [band pwr mkr on], [power ratio C/N]. 6 Press [Marker|Entry] Rotate the knob to move the measurement band from the signal to a noise area. The display should appear as below. The grid lines have been turned off to highlight the band power markers.

The diamond shaped marker provides a reference point
Carrier to noise ratio
The value indicated in the lower left corner of the display reflects the difference between the marker level at the carrier peak and the total noise within the band markers.
Measured noise band

Making Simple Noise Measurements
7 Change to a normalized noise measurement: Toggle to [power ratio C/No] The display should appear as below. The grid lines have been turned off to highlight the band power markers.
The carrier-to-noise and carrier-to-normalized-noise marker measurements require that the standard (diamond shaped) marker be on the signal peak as a reference. If the marker is not on, the displayed value will only reflect the noise level.
Step 3 above illustrates that there are two ways to perform certain actions—by using the hardkey/softkey sequence or by using the short-cut shift/hardkey sequence.

The diamond shaped marker provides a reference point
Carrier-to-noise ratio normalized to one Hertz
Now the value indicated in the lower left corner of the display reflects the difference between the marker level at the carrier peak and the noise-power within the band markers normalized to one Hertz bandwidth.
You can perform band power measurements in either Vector or Scalar Mode. If you use Scalar mode and you have selected a combination of resolution bandwidth, window type, and number of frequency points such that the analyzer implements the detector, the analyzer will prompt you to select the sample detector in order to calculate the band power accurately.
Measured noise band



This chapter uses the time gating feature to analyze a multi-burst signal which is provided on the Signals Disk which accompanies the analyzer’s Operator’s Guide. Time gating allows you to isolate a portion of a time record for further viewing and analysis. For more details on time gating concepts see “Gating Concepts” in the Operator’s Guide.

Using Gating to Characterize a Burst Signal
First we’ll look at the spectrum of the signal and see that three components exist. Then we’ll look at the time display of the burst signal and analyze each burst separately to determine which spectral components exist in each burst.
1 Select the baseband receiver mode and initialize the analyzer: Press [Instrument Mode] [receiver] [RF section (0-10 MHz)]. Press [Preset]. 2 Load the source signal file BURST.DAT into data register D3: Insert the Signals Disk in the analyzer’s disk drive. Press [Save/Recall], [default disk], [internal disk] to select the internal disk drive. Press [Return] (bottom softkey), [catalog on] to display the files on the disk. Rotate the knob until the file BURST.DAT is highlighted. Press [recall trace], [from file into D3], [enter]. 3 Connect the SOURCE output to the INPUT with a BNC cable. 4 Turn on the source and select arbitrary signal D3: Press [Source], [source on], [source type], [arb data reg], [D3], [Return], [arbitrary]. Press [Average], [average on]. The display should now appear as shown below.

The spectrum with averaging turned on. Note existence of three components.

Using Gating to Characterize a Burst Signal
5 Configure the display and the measurement: Press [Display], [2 grids], [more display setup], [grids off]. Press [B], [Measurement Data], [main time] (toggle to ch1 on a 2-channel analyzer). Press [Ref Lvl/Scale], [Y per div], 50, [mV]. Press [Trigger], [trigger type], [internal source]. Press [Time], [main length], 32, [us]. 6 Set up the time gating and examine the first burst: Press [Time], [gate on], [gate length], 10, [us]. Press [ch1 gate dly], [Marker|Entry] Rotate the knob until the gate is at each end of the first burst signal. The display should now appear as shown to the left below. 7 Examine the second burst: Rotate the knob until the gate is at each end of the second burst signal. The display should now appear as shown to the right below.


Spectrum (top trace) of the burst is derived by gating the time signal (bottom trace). The gate’s delay and length are selected to encompass the burst signal (vertical markers show gate position). Note existence of the first spectral component in the left display and the existence of the other two components in the right display.



This section shows you how to make typical relative phase measurements on modulated carrier signals. In this example, you measure the phase of sidebands on AM and PM signals relative to the carrier. The test signals are provided on the Signals Disk which accompanies the analyzer’s Operator’s Guide.

1 Select the baseband receiver mode and initialize the analyzer: Press [Instrument Mode] [receiver] [RF section (0-10 MHz)]. Press [Preset]. 2 Load AM and PM signals from the Signals Disk into registers and play the AM signal through the source: Insert the Signals Disk in the internal disk drive. Use the BNC cable to connect the SOURCE output to the INPUT. Press [Save/Recall], [default disk], [internal disk]. Press [Return], [catalog on]. Rotate the knob to highlight AMSIG.DAT Press [recall trace], [from file into D1], [enter]. Rotate the knob to highlight PMSIG.DAT Press [from file into D2] [enter]. Press [Source], [source on], [source type], [arbitrary]. 3 Configure the measurement and display: Press [Frequency], [span], 150, [kHz], Press [Trigger], [trigger type], [internal source], Press [Sweep], [single]. 4 Activate a different trace as a phase display: Press [Display], [2 grids], Press [B], [Measurement Data], [spectrum] (select ch1 with a 2-channel analyzer), Press [Data Format], [phase wrap] 5 Start a single sweep: Press [Pause|Single].
Measuring Relative Phase

6 Activate two traces: Press [Shift], [A] (two Active Trace LEDs are now turned on) 7 Turn on marker coupling and zero the offset marker on the carrier: Press [Marker], [couple mkrs on], Press [Shift], [Marker] to place the marker on the carrier peak, Press [Shift], [Marker⇒] to zero the offset marker. 8 Use the search marker to measure the phase of the two largest sidebands relative to the carrier: Press [Marker Search], [next peak], and note the phase displayed for the lower trace. Press [next peak] again and note the phase.

The phase values vary with each sweep but for an AM signal the average phase of the sidebands is equal to the carrier phase.
Measuring Relative Phase

Continue from “To measure the relative phase of an AM signal.” 1 Replace the arbitrary source AM signal with the PM signal in register D2: Press [Source], [source type], [arb data reg], [D2]. 2 Start a single sweep: Press [Pause|Single]. 3 Zero the offset marker on the carrier: Press [Shift], [Marker], Press [Shift], [Marker⇒] 4 Use the search marker to measure the phase of the two largest sidebands relative to the carrier: Press [Marker Search], [next peak] and note the phase displayed for the lower trace. Press [next peak] again and note the phase.

The phase values vary with each sweep but for a PM signal the average phase of the two sidebands is equal to –90 degrees from the carrier.
Measuring Relative Phase

This section shows you how to make a typical network measurement. In this example, we will be characterizing a 4.5 MHz bandpass filter.

You must use the source output and the channel 1 and channel 2 inputs on the IF section for network measurements.
1 Using a BNC “T” adapter or power splitter and BNC cables, connect the analyzer’s SOURCE to the CHANNEL 1 input directly and to the CHANNEL 2 input through a filter as shown in the illustration below. 2 Select the IF baseband receiver mode and initialize the analyzer: Press [Instrument Mode] [receiver] [IF section (0-10 MHz)]. Press [Preset]. 3 Configure the analyzer to make two-channel frequency response measurements: Press [Measurement Data], [freq response].
Characterizing a Filter

4 Configure the source and measurement for a frequency response measurement: Press [Source], [source on], Press [source type], [periodic chirp], Press [Return], (bottom softkey) Press [level], .5, [ Vrms].
Press [Res Bw/Window], [rbw mode arb], Press [main window], [uniform].
Press [Range], [channel both], [ch* single range up-down].
Press [Average], [average on] Press [num averages], 50, [enter], Press [average type], [rms (video)].
Press [Auto Scale]. 5 Start an averaged measurement: Press [Meas Restart].
The display should appear similar to that shown below. To learn more about the choices you make in this measurement, display online help for the various keys used (see “Using Online Help” if you are not familiar with how to do this).

Frequency response data displays the output of a device-under-test divided by the input
Characterizing a Filter

Continue from “To set up a frequency response measurement.” 1 Move the marker to the largest part of the frequency response trace: Press [Marker⇒], [marker to peak]. or Press [Shift], [Marker] 2 Move the marker with the knob to view the absolute gain/loss of this particular filter network at different frequencies.

The frequency and amplitude of the trace at the marker location are shown at the top of the display
Characterizing a Filter

Continue from “To set up a frequency response measurement” or from “Using the absolute marker.” 1 Move the marker to the largest part of the frequency response trace if it is not already there: Press [Shift], [Marker]. 2 Establish the reference point for the relative (offset) marker: Press [Marker], [zero offset] or Press [Shift], [Marker⇒] 3 Move the marker with the knob to view the relative gain/loss of this particular filter at different frequencies. The offset marker allows you to establish a reference point with the square-shaped marker. As you move diamond-shaped marker, the value displayed by the marker readout reflects the difference between the reference point and the marker.

The marker frequency and amplitude reflect the value of the diamond-shaped marker relative to the offset (square) marker
Characterizing a Filter

Complete “To set up a frequency response measurement” or continue from one of the previous marker measurements. 1 Move the marker to the largest part of the frequency response trace if it is not already there: Press [Shift], [Marker]. 2 Activate and zero the offset marker if it is not already activated: Press [Shift], [Marker⇒]. 3 Define the search target level and perform a search: Press [Marker Search], [search setup], Press [search target], −6, [dB], Press [search right], [search left]. The search marker allows you to quickly find a target value. When the offset marker is activated the target value is relative to the reference point.

With the offset marker activated, the search marker indicates the point on the trace which is separated from the offset marker by the target value
The search marker finds a Y-axis value with reference to a target value
Characterizing a Filter

Complete “To set up a frequency response measurement” or continue from one of the previous marker measurements. 1 Display a second trace: Press [Display], [2 grids]. 2 Activate the second trace and define it as a frequency response measurement: Press [B], [Measurement Data], [frequency response]. 3 Specify phase data for the second trace: Press [Data Format], [phase wrap]. 4 Couple the markers on traces A and B: Press [Marker], [couple mkrs on]. 5 Move the markers with the knob to determine phase with respect to frequency response. 6 Overlap the two traces: Press [Shift], [A]. Press [Display] [single grid].
In this example, note that a trace which is displayed is not necessarily active (capable of being configured). You must specifically activate a displayed trace in order to change its configuration. For example, if you have chosen the relative marker in one trace then couple the markers, the marker on the second trace will be absolute, rather than relative, unless you activate the second trace and select the relative marker.

Coupling the markers on two traces lets you compare values at the same frequency
Characterizing a Filter

Complete “To set up a frequency response measurement” or continue from one of the previous measurements. 1 Display a second trace: Press [Display], [2 grids]. 2 Activate the second trace and select a coherence measurement: Press [B], [Data Format], [magnitude linear], [Measurement Data], [more choices], [coherence].

Characterizing a Filter

This chapter shows you how to perform various common tasks. These include setting up and using peripherals and defining and using math functions.

To set up peripherals.
You may connect peripherals to three ports—one GPIB port, one serial port, and one parallel port. GPIB peripherals may include printers, plotters, and external disk drives. Supported serial devices are plotters and printers. Certain printers are parallel devices. 1 Connect the ports of your peripheral and analyzer with the correct cables. See “Preparing the Analyzer for Use” for information on physical connections. 2 Turn on the peripherals. 3 Set up GPIB peripherals: Determine the address of the peripheral from your peripheral’s documentation Use this as below. On the analyzer, press [Local/setup], [peripheral addresses]. Press the softkey corresponding to your device type. Press , [enter]. Repeat this step for each GPIB peripheral. 4 Set up serial peripherals: Refer to your serial device’s documentation to select correct setup parameters. Press [Serial 1 setup] and enter the correct parameters.
Display online help for more details on setup and parameter choices.

General Tasks

1 Set up your printer or plotter if you haven’t already done so. 2 Select the output format and device type: Press [Plot/Print], [output fmt] and select the desired format. Press [device defaults] and select a device if you want other than the default. 3 Select the type of output port: Press [Plot/Print], [output to] and select the port to which your printer or plotter is attached. 4 Press [Local/Setup], [system controller]. 5 Press [Plot/Print], [start plot/print]
The analyzer is only able to initiate printing or plotting if it is attached to a printer or plotter and is designated as the system controller. If you haven’t already set up your printer or plotter, see “To set up peripherals.” All of the screen’s contents, except the softkey labels, are printed when you complete this task.
You may select various parameters under the [plot item] and [plot/print setup] softkeys depending on your particular peripheral. To learn more about these parameters, display online help for the relevant softkeys.

With a plotter, you can elect to plot portions of the display
You can control certain device options, depending on your output device
You must use the PLOT/PRINT menu to select the correct type of device and port before starting a plot or print of the screen contents.
General Tasks

You may save trace data, instrument states, trace math functions, instrument BASIC programs, and time-capture buffers. 1 Select the default disk: Press [Save/Recall], [default disk] Press [nonvolatile RAM disk], [volatile RAM disk] or [internal disk] 2 Press [Return]. 3 Press the softkey that matches the type of data you want to save. 4 Enter the file name if you have chosen to save to a file: Use the hardkeys (which have now been remapped to represent the symbols etched to the lower right of them), softkeys, knob, and numeric keys to type in a file name. 5 Press [enter].
For more information on the softkeys and parameter choices, display online help.
If you are using the internal disk drive, you must insert a formatted 3.5-inch flexible disk into the analyzer’s internal disk drive. If you want to save data but the disk has not been previously formatted see “To format a disk.”
General Tasks

You may recall trace data, instrument states, trace math functions, instrument BASIC programs, and time-capture buffers. 1 Select the default disk: Press [Save/Recall], [default disk] Press [nonvolatile RAM disk], [volatile RAM disk] or [internal disk] 2 Press [Return]. 3 To easily recall a file you may press [catalog on] to display the names of files stored on the disk then use the knob to scroll to the desired file. 4 Press the softkey that matches the type of data you want to recall (then select a storage register if you are recalling a trace). 5 If you have not selected a file name from the catalog, enter the file name: Use the hardkeys (which have now been remapped to represent the symbols etched to the lower right of them), softkeys, knob, and numeric keys to type in a file name. 6 Press [enter].
For more information on the softkeys and parameter choices, display online help.
General Tasks

1 Select the disk drive you want to format: Press [Disk Utility], [default disk]. Press the softkey corresponding to the disk drive you want to format. 2 Press [Return], [format disk]. Select appropriate parameters for your disk drive (disk type, interleave etc.). 3 Press [perform format], [proceed].
You may format 3.5-inch disks in the internal disk drive. They must be double-sided, high-density flexible disks that are not write-protected.
The analyzer may take a few minutes to format a disk (depending on the type of disk) and is unavailable for other tasks during that time.
General Tasks

In this section you learn how to create a math function which inverts a signal. 1 Initialize the analyzer: Press [Preset] 2 Define a constant: Press [Math], [define constant], [define K1] Press [real part], 1, [enter], [imag part], 0, [enter]. 3 Define a math function: Press [Math], [define F1] Press [constant], [K1], [/], [meas data], [spectrum], [enter]. A math function remains in memory through a Preset but will be erased when you power down the analyzer. If you want to preserve the math function for future use, save it in the non-volatile RAM or on an internal disk.

General Tasks

In this section you learn how to apply a a math function to a signal. This task assumes that you have completed “To create a math function.” 1 Initialize the analyzer: Press [Preset] 2 Provide an averaged signal from the internal source: Press [Source], [source on], [Average], [average on]. 3 Apply the inversion math function you created to this signal: Press [Measurement Data], [math func], [F1]. 4 Press [Auto Scale].

General Tasks

1 Press [View State]. 2 Press [measurement state] or [input/source state]. These summaries reflect the current states of important measurement, input, and source parameters. You may use these summaries to:
l quickly check the current setup
l document the setup (The list can be printed or plotted.)
You will note that the contents of the measurement state differ depending on the instrument mode. This reflects the fact that some parameters are not used for a particular instrument mode.


General Tasks




This chapter contains instructions for inspecting and installing the analyzer. This chapter also includes instructions for cleaning the screen, transporting and storing the analyzer.
Power Requirements The analyzer can operate from a single-phase ac power source supplying voltages as shown in the table. With all options installed, the total power consumption of both sections is less than 1025 VA.
AC Line Voltage Range Frequency 90-140 Vrms 47-63 Hz 198-264 Vrms 47-63 Hz
The line-voltage selector switches are set at the factory to match the most commonly used line voltage in the country of destination; the appropriate fuses are also installed. To check or change either the line-voltage selector switch or the fuse, see the appropriate sections later in this chapter.

Preparing the Analyzer for Use
Power Cable and Grounding Requirements On the GPIB connector, pin 12 and pins 18 through 24 are tied to chassis ground and the GPIB cable shield. The instrument frame, chassis, covers, and all exposed metal surfaces including the connectors’ outer shell are connected to chassis ground. However, if channel 2 in the IF section is not installed, the channel 2 BNC connector’s outer shell is not connected to chassis ground.
The analyzer is equipped with two three-conductor power cords which ground the analyzer when plugged into appropriate receptacles. The type of power cable plug shipped with each analyzer depends on the country of destination. The following figure shows available power cables and plug configurations.

Preparing the Analyzer for Use
*The number shown for the plug is the industry identifier for the plug only, the number shown for the cable is an Agilent part number for a complete cable including the plug.
**UL listed for use in the United States of America. Warning The power cable plug must be inserted into an outlet provided with a protective earth terminal. Defeating the protection of the grounded analyzer cabinet can subject the operator to lethal voltages.

Preparing the Analyzer for Use
The analyzer was carefully inspected both mechanically and electrically before shipment. It should be free of marks or scratches, and it should meet its published specifications upon receipt. 1 Inspect the analyzer for physical damage incurred in transit. If the analyzer was damaged in transit, do the following:
l Save all packing materials.
l File a claim with the carrier.
l Call your Agilent Technologies sales and service office.
2 Check that the line-voltage selector switches are set for the local line voltage.
The line-voltage selector switches are set at the factory to match the most commonly used line voltage in the country of destination. To check or change the line-voltage selector switches, see ‘’To change the IF section’s line-voltage switch’’ and ‘’To change the RF section’s line-voltage switch.’’ 3 Check that the correct line fuses are installed in the fuse holders. The fuses are installed at the factory for the most commonly used line voltage in the country of destination. There is one line fuse in the IF section and one line fuse in the RF section. To determine if the correct line fuses are installed, see ‘’To change the IF section’s fuse’’ and ‘’To change the RF section’s fuse.’’ 4 Connect the IF section to the RF section. For instructions on connecting the sections, see ‘’To connect the sections.’’ 5 Using the supplied power cords, plug the analyzer’s IF section and RF section into appropriate receptacles. The analyzer is shipped with two three-conductor power cords that ground the analyzer when plugged into appropriate receptacles. The type of power cable plug shipped with each analyzer depends on the country of destination.

Preparing the Analyzer for Use
6 Set the RF section’s rear panel and front panel power switches to on. Press the ‘’l‘’ symbol end of the rocker-switches located on the lower right of the rear panel and on the lower left of the front panel. The RF section provides standby power for the high precision frequency reference. The rear-panel line switch interrupts all power including standby power when you press the ‘’O‘’ symbol end of the switch. The front-panel power switch interrupts all power except standby power when you press the ‘’OI ‘’ symbol end of the switch. 7 Set the IF section’s power switch to on. Press the ‘’l‘’ symbol end of the rocker-switch located on the lower left of the front panel. The analyzer requires about 30 seconds to complete its power-on routine. 8 Test the electrical performance of the analyzer using the operation verification or the performance tests in chapter 2, ‘’Verifying Specifications’’ in the Installation and Verification Guide. The operation verification tests verify the basic operating integrity of the analyzer; these tests take about 2.5 hours to complete and are a subset of the performance tests. The performance tests verify that the analyzer meets all the performance specifications; these tests take about 5 hours to complete.

Preparing the Analyzer for Use
Do NOT use the IF section’s EXT REF OUT connector or optional OVEN REF OUT connector as an external reference output. 1 Attach the IF section to the RF section.
If the hardware is not installed, follow the instructions supplied with the Rear Panel Lock Foot Kit. If the hardware is already installed, slide the IF section on top of the RF section making sure the front lock-links engage the IF section’s frame. Screw the rear lock feet together. 2 Connect the RF section’s SERIAL 2 port to the IF section’s SERIAL 2 port using the supplied serial interface interconnect cable. Make sure the end of the cable with the EMI suppressor is conected to the IF section. 3 Connect the RF section’s OVEN REF OUT connector to the EXT REF IN connector using the supplied coax BNC-to-coax BNC connector. If the RF section does not have the OVEN REF OUT connector (option AY4, Delete High Precision Frequency Reference), connect a 1 MHz, 2 MHz, 5 MHz, or 10 MHz sine or square wave, with an amplitude greater than 0 dBm to the RF section’s EXT REF IN connector. For best residual phase-noise, use 10 MHz with an amplitude greater than or equal to 5 dBm. See the Technical Data publication in the beginning of your Installation and Verification Guide for specifications that require the high precision frequency reference. 4 Connect the RF section’s 10 MHz REF TO IF SECTION connector to the IF section’s EXT REF IN connector using the supplied 12-inch BNC-to-BNC cable.


Preparing the Analyzer for Use
5 Connect the IF section’s SOURCE connector to the RF section’s IN connector using the supplied 8.5-inch BNC-to-BNC cable. 6 Connect the IF section’s CHANNEL 1 connector to the RF section’s OUT connector using the supplied 8.5-inch BNC-to-BNC cable.

Preparing the Analyzer for Use
The analyzer is shipped with plastic feet in place, ready for use as a portable bench analyzer. The plastic feet are shaped to make full-width modular instruments self-align when they are stacked.
l Install the analyzer to allow free circulation of cooling air.
Cooling air enters the analyzer through the rear panel and exhausts through both sides.
l Protect the analyzer from moisture and temperatures or temperature changes
that cause condensation within the analyzer. The operating environment specifications for the analyzer are listed in the Technical Data publication in the beginning of your Installation and Verification Guide.
l To install the analyzer in an equipment cabinet, follow the instructions shipped
with the rack mount kits.

Preparing the Analyzer for Use
The line-voltage selector switch is set at the factory to match the most commonly used line voltage in the country of destination. 1 Unplug the power cord from the IF section (the section with ‘’Agilent 89431A’’ silk screened on the lower right rear panel). 2 Slide the line voltage selector switch to the proper setting for the local line voltage. 3 Check to see that the proper fuse is installed. See ‘’To change the IF section’s fuse.’’


90-140 Vrms 47-440 Hz 115 198-264 Vrms 47-63 Hz 230

Preparing the Analyzer for Use
The line-voltage selector switch is set at the factory to match the most commonly used line voltage in the country of destination. 1 Unplug the power cord from the RF section (the section with “Agilent 89431A” silk screened on its lower left rear panel). 2 Using a small screw driver, pry open the power selector cover.
3 Remove the cylindrical line voltage selector. 4 Position the cylindrical line voltage selector so the required voltage will be facing out of the power selector, then reinstall.
Selector Switch Range Frequency
90-110 Vrms 47-63 Hz 100 103-140 Vrms 47-63 Hz 120 198-242 Vrms 47-63 Hz 220 216-264 Vrms 47-63 Hz 240
5 Check to see that the proper fuse is installed. See ‘’To change the RF section’s fuse.’’ 6 Close the power selector by pushing firmly on the power selector cover. 7 Check that the correct line voltage appears through the power selector cover.

Preparing the Analyzer for Use
The fuse is installed at the factory to match the most commonly used line voltage in the country of destination. 1 Unplug the power cord from the IF section (the section with ‘’Agilent 89410A’’ silk screened on its lower right rear panel). 2 Using a small screw driver, press in and turn the fuse holder cap counter-clockwise. Remove when the fuse cap is free from the housing.

3 Pull the fuse from the fuse holder cap. 4 To reinstall, select the proper fuse and place in the fuse holder cap.

90-140 Vrms 47-440 Hz 115 2110-0342 8 A 250 V Normal Blow 198-264 Vrms 47-63 Hz 230 2110-0055 4 A 250 V Normal Blow
5 Place the fuse holder cap in the housing and turn clockwise while pressing in.

Preparing the Analyzer for Use
The fuse is installed at the factory to match the most commonly used line voltage in the country of destination. 1 Unplug the power cord from the RF section (the section with “Agilent 89431A” silk screened on its lower left rear panel). 2 Using a small screw driver, pry open the power selector cover.

3 Pull the white fuse holder out of the power selector and remove the fuse from the fuse holder. 4 Select the proper fuse and place in the fuse holder.

Agilent 89431A 90-110 Vrms 47-63 Hz 100 2110-0381 3 A 250 V Slow Blow Agilent 89431A 103-140 Vrms 47-63 Hz 120 2110-0381 3 A 250 V Slow Blow Agilent 89431A 198-242 Vrms 47-63 Hz 220 2110-0304 1.5 A 250 V Slow Blow Agilent 89431A 216-264 Vrms 47-63 Hz 240 2110-0304 1.5 A 250 V Slow Blow
1 Align the white arrow on top of the fuse holder with the white arrow on the power selector cover. All three arrows should point in the same direction. Push the fuse holder into the top slot of the power selector. 2 Close the power selector by pushing firmly on the power selector cover. 3 Check that the correct line voltage appears through the power selector cover.

Preparing the Analyzer for Use
Analyzers with option UFG, 4 megabyte extended RAM and additional I/O, have a ThinLAN and AUI (attachment unit interface) port for connecting the analyzer to the LAN (local area network). 1 Set the power switch to off ( O ).
2 Connect the ThinLAN BNC cable to the ThinLAN port or the appropriate media access unit (MAU) to the AUI port. 3 Set the power switch to on ( l ). 4 Press the following keys: [Local/Setup]
[LAN port setup] [port select ThinLAN (BNC)] or [port select AUI (MAU)] [IP address] internet protocol address [Return]
[LAN power-on active]
See your LAN system administrator for the internet protocol address. Your LAN system administrator can also tell you if you need to set the gateway address or subnet mask.

Preparing the Analyzer for Use
The IF section’s Serial 1 port is a 9-pin, EIA-574 port that can interface with a printer or plotter. The total allowable transmission path length is 15 meters.
l Connect the IF section’s SERIAL 1 port to a printer or plotter using a 9-pin
female to 25-pin RS-232-C cable.
Part Number Cable Description
Agilent 24542G 9-pin female EIA-574 to 25-pin male RS-232 HP 24542H 9-pin female EIA-574 to 25-pin female RS-232
The IF section’s Parallel Port is a 25-pin, Centronics port. The Parallel Port can interface with PCL printers or HP-GL plotters.
l Connect the IF section’s rear panel PARALLEL PORT connector to a plotter or
printer using a Centronics interface cable.

Preparing the Analyzer for Use
The analyzer is compatible with the General Purpose Interface Bus (GPIB). Total allowable transmission path length is 2 meters times the number of devices or 20 meters, whichever is less. Operating distances can be extended using an GPIB Extender.
Analyzers with option UFG, 4 megabytes extended RAM and additional I/O, have an additional GPIB connector. The additional GPIB connector, SYSTEM INTERCONNECT, is only for connection to the spectrum analyzer used with the Agilent 89411A 21.4 MHz Down Converter.
l Connect the analyzer’s rear panel GPIB connector to an GPIB device using an
GPIB interface cable.
For GPIB programming information, see the Agilent 89400 Series GPIB Command Reference.
The External Monitor connector is a 15-pin connector with standard VGA pinout. The External Monitor connector can interface with an external, multi-scanning monitor. The monitor must have a 25.5 kHz horizontal scan rate, a 60 Hz vertical refresh rate, and must conform to EIA-343-A standards.
l Connect the analyzer’s rear panel EXTERNAL MONITOR connector to an
external monitor using an appropriate cable. For additional information, see ‘’EXTERNAL MONITOR connector’’ in the analyzer’s online help.

Preparing the Analyzer for Use
The analyzer may be connected to an optional external keyboard. The keyboard remains active even when the analyzer is not in alpha entry mode. This means that you can operate the analyzer using the external keyboard rather than the front panel. Pressing the appropriate keyboard key does the same thing as pressing a hardkey or a softkey on the analyzer’s front panel. 1 Set the IF section’s power switch to on ( l ). 2 Connect the round plug on the keyboard cable to the KEYBOARD connector on the analyzer’s front panel. Make sure to align the plug with the connector pins.

3 Connect the other end of the keyboard cable to the keyboard.

Preparing the Analyzer for Use
4 To configure your analyzer for a keyboard other than U.S. English, press [System Utility] [keyboard type]. Then press the appropriate softkey to select the language. Configuring your analyzer to use a keyboard other than U.S. English only ensures that the analyzer recognizes the proper keys for that particular keyboard. Configuring your analyzer to use another keyboard does not localize the on-screen annotation or the analyzer’s online HELP facility.
The minimum loss pad (option 1D7) provides a 50 ohm matched impedance to the analyzer and a 75 Ωmatched impedance to the device under test. 1 Connect the minumum loss pad to the RF section’s INPUT or SOURCE connector. 2 Connect a 75 Ωcable between the minimum loss pad and the device under test. Use either a 75 Ωtype-N cable or the supplied 75 Ωtype-N(m)-to-BNC(f) adapter and a 75 ΩBNC cable.

Preparing the Analyzer for Use
The analyzer screen is covered with a plastic diffuser screen (this is not removable by the operator). Under normal operating conditions, the only cleaning required will be an occasional dusting. However, if a foreign material adheres itself to the screen, do the following: 1 Set the IF section’s power switch to off ( O ).
2 Remove the power cord. 3 Dampen a soft, lint-free cloth with a mild detergent mixed in water. 4 Carefully wipe the screen.
To prevent damage to the screen, do not use cleaning solutions other than the above.
l Store the analyzer in a clean, dry, and static free environment.
For other requirements, see environmental specifications in the Technical Data publication in the beginning of your Installation and Verification Guide.

Preparing the Analyzer for Use
l Disconnect the IF section from the RF section and package each section using
the original factory packaging or packaging identical to the factory packaging. Containers and materials identical to those used in factory packaging are available through Hewlett-Packard offices.
l If returning the analyzer to Hewlett-Packard for service, attach a tag to each
container describing the following:
l Type of service required
l Full serial number In any correspondence, refer to the analyzer by model number and both serial numbers.
l Mark the containers FRAGILE to ensure careful handling.
l If necessary to package the analyzer in containers other than original
packaging, observe the following (use of other packaging is not recommended):
l Wrap each section in heavy paper or anti-static plastic.
l Protect the front panels with cardboard.
l Use double-wall cartons made of at least 350-pound test material.
l Cushion each section to prevent damage.
l Return address · l Model number

Preparing the Analyzer for Use
Check that the power cord is connected to the IF section and to a live power
source.
Check that the front-panel switch is on ( l ).
Check that the voltage selector switch is set properly.
See ‘’To change the IF section’s line-voltage switch’’ on page 7-10.
Check that the fuse is good.
See ‘’To change the IF section’s fuse’’ on page 7-12.
Check that the IF section’s air circulation is not blocked.
Cooling air enters the IF section through the rear panel and exhausts through both sides. If the IF section’s air circulation is blocked, the IF section powers down to prevent damage from excessive temperatures. The IF section remains off until it cools down and its power switch is set to off ( O ) then to on ( l ).
Obtain service, if necessary. See ‘’Need Assistance?’’ at the end of this guide.

Preparing the Analyzer for Use
Check that the power cord is connected to the RF section and to a live power
source.
Check that the RF section’s rear panel and front panel power switches are on
( l ).
Check that the voltage selector switch is set properly.
See ‘’To change the RF section’s line-voltage switch’’ on page 7-11.
Check that the fuse is good.
See ‘’To change the RF section’s fuse’’ on page 7-13.
Check that the RF section’s air circulation is not blocked.
Cooling air enters the RF section through the rear panel and exhausts through both sides. If the RF section’s air circulation is blocked, the RF section powers down to prevent damage from excessive temperatures. The RF section turns back on when it cools down.
this guide.

Preparing the Analyzer for Use
Check that the RF section’s fan is running.
If the fan is not running, see ‘’If the RF section will not power up.’’
Check that the Serial 2 port on the IF section and on the RF section are
connected together.
Press [Instrument Mode] and check that the receiver softkey displays ‘’RF section
(2-2650 MHz)’’. If the receiver softkey does not display ‘’RF section (2-2650 MHz)’’ press [receiver] [ RF section (2-2650 MHz) ].
Leaving the RF section on, turn the IF section off ( O ) then on ( l ).
The IF section will not detect the RF section if the RF section was not on before the IF section performs the power-on routine.
Obtain service, if necessary. See ‘’Need Assistance?’’ at the end of this guide.



block diagrams (continued) connectors, rear panel HT digital demodulation OP 17-3, HT FFT overview HT general OP 13-11 IBASIC memory HT instantaneous spectrums HT main memory HT measurement data HT measurement memory HT scalar OP 12-3, HT source block diagram HT time capture HT traces HT vector OP 12-5, HT video demodulation OP 18-3 block size
See information bandwidth See noise equivalent bandwidth See resolution bandwidth See window bandwidth bandwidth coupling OP 14-4 baseband measurements OP 14-3, HT baseband signals
digital demod OP 6-12 video demod OP 7-15 BASIC HT about IBASIC HT available utilities HT creating a program with keystrokes HT debugging a program HT deleting a function HT deleting a program HT deleting a subprogram HT deleting multiple functions HT deleting multiple subprograms HT displaying a program HT displaying graphics HT editing a program HT executing a single program line HT indenting nested programs HT memory allocations HT printing a program HT program error messages HT program variables HT re-saving programs HT recalling programs HT renumbering a program HT resetting a program HT running a program HT saving programs HT securing a program HT specifying columns HT stack space, auto allocation HT stack space, manual allocation HT stopping a program HT beeper on/off softkey HT bins
See time record size BPSK/8PSK softkey HT BT, setting HT burst modulation
about calibration HT adjusting calibration data HT analog demodulation, calibrating HT auto zero calibration HT auto zero calibration, single HT performing a single calibration HT saving calibration data HT time-domain calibration HT capture
See time capture carrier frequency error
(in symbol table) HT carrier locking
analog demod OP 15-8 video demod OP 17-8, OP 18-8 carrier offset (FSK) HT carrier to noise markers GS 2-4 carrier, auto OP 15-8 center frequency HT setting center frequency HT setting with the marker HT signal tracking HT centronics port GS 7-15
See frequency points bins, defined HT bit patterns, digital demodulation
See symbol states block diagrams
analog demodulation OP 15-3, HT arbitrary source HT ch1 + j*ch2 receiver HT connectors, front panel HT
GS = Getting Started Guide HT = Online Help OP = Operator's Guide (press (Help) key)
Index · bandwidth · See demodulation, digital

ch1 + j*ch2 receiver HT changing numeric parameters HT channel 1
cross spectrum measurements HT CW (fixed sine) softkey HT
D data comment header HT data format hardkey HT data registers HT about data registers HT displaying data-register contents HT removing HT DATA?, definition of HT date, changing HT dc measurements OP 14-9 dc offset softkey HT dc response OP 14-9 decibels HT DECT (digital european
See input channels channel 1 and 2 connectors HT channel 2
See input channels circulation, air GS 7-9 cleaning the screen GS 7-19 clear source trip softkey HT coherence GS 5-8, HT color index softkey HT color, adjusting HT connecting
frequency reference GS 7-7 IF section to RF section GS 7-7 minimum loss pad GS 7-18 connector
AUI GS 7-14 external monitor GS 7-16 external reference GS 7-7 grounding requirements GS 7-3 GPIB GS 7-16 input GS 7-8 keyboard GS 7-17 oven reference GS 7-7 parallel GS 7-15 serial GS 7-15 source GS 7-8 system interconnect GS 7-16 ThinLAN GS 7-14 connectors, front panel connectors HT connectors, rear panel connectors HT constellation diagram
demodulation, analog HT about analog demodulation HT affects on bandwidth HT AM demodulation HT auto carrier HT averaging OP 15-13 block diagram OP 15-3, HT carrier frequency, displaying HT carrier locking OP 15-8 examples OP 1-1 FM demodulation HT gating OP 15-13, OP 16-5 making zero span measurements HT overview OP 12-6 PM demodulation HT triggering OP 15-13, HT two channel measurements HT See also AM, FM, PM demodulation
example OP 6-5, OP 7-12 using HT continuing a measurement HT continuous sweep, selecting HT cooling GS 7-9 coordinates, trace
demodulation, digital HT about digital demodulation HT aliasing, IQ measured spectrum HT amplitude droop (in symbol table) HT averaging HT block diagrams OP 17-3, HT BPSK/8PSK HT carrier frequency error
See traces copying files between 3.5 inch diskettes HT cords, power GS 7-3 correlation HT
auto correlation HT cross correlation softkey HT cross-correlation, math function HT coupling
(in symbol table) HT carrier locking HT connections HT constellation diagram HT
GS = Getting Started Guide HT = Online Help OP = Operator's Guide (press (Help) key)
Index · cellular telephone) HT · bandwidth OP 14-4

demodulation, video HT about video demodulation HT block diagram OP 18-3 capabilities OP 18-2 carrier frequency error (in symbol table) HT carrier locking OP 17-8, OP 18-8, HT center frequency, setting OP 7-4 constellation diagram HT constellation points, viewing OP 8-5 DVB QAM 16/32/64 HT DVB QAM state definitions HT DVB QAM, example OP 7-10 error-vector magnitude (in symbol table) HT example OP 8-1 eye diagram HT filtering HT filters, user defined OP 9-6 formatting and displaying OP 18-7 formatting displays OP 7-12 hexadecimal, in symbol table HT I/Q measured, displaying HT I/Q origin offset (in symbol table) HT ideal states HT magnitude error HT maximum time points, setting HT mirrored spectrums OP 18-17, HT normalization HT phase error (in symbol table) HT points-per-symbol, setting HT QAM 16/32/64/256 HT range, setting OP 18-10 setup OP 7-2 setup, 32 DVB QAM OP 7-8 signal creation OP 9-2 SNR HT state definitions, defining HT state definitions, displaying HT symbol clock, adjusting HT symbol locking OP 17-8, OP 18-8 symbol rate, setting HT sync patterns, pre-defined HT sync search OP 7-13 trellis diagram HT troubleshooting HT VSB 8/16 HT VSB, example OP 7-6
constellation points, viewing OP 8-5 DQPSK HT error-vector magnitude (in symbol table) HT error-vector magnitude trace HT example OP 8-1 eye diagram HT filtering OP 17-16, HT filters, user defined OP 9-6 formatting displays OP 17-7 I/Q measured, displaying HT I/Q origin offset (in symbol table) HT I/Q reference, displaying HT ideal states HT magnitude error HT magnitude error, FSK HT magnitude-error trace HT maximum time points, setting HT memory allocation HT MSK (minimum shift keying) HT normalization HT phase error (in symbol table) HT phase-error trace HT points-per-symbol, setting HT power calculation OP 8-4 pulse modulation HT pulse search OP 6-6 QAM HT QPSK (quadrature psk) HT setting up HT setup OP 6-2, OP 17-6 signal creation OP 9-2 SNR HT span, for spectrum displays HT standard setups, example OP 6-4, HT state definitions, defining HT state definitions, displaying HT symbol clock, adjusting HT symbol rate, setting HT symbol table HT sync offset, setting HT sync pattern, setting HT sync patterns, pre-defined HT sync search OP 6-8 sync words, using HT tips for using digital demodulation HT trellis diagram HT troubleshooting HT See also FSK
GS = Getting Started Guide HT = Online Help OP = Operator's Guide (press (Help) key)
Index · demodulation, digital (continued)

detection OP 14-12, HT
equivalent noise bandwidth HT error & status messages HT error analysis
about detection HT normal detector HT positive peak detector HT sample detector HT deviation (FSK) HT diagnostics softkey HT digital demodulation
digital demod OP 8-12 video demod OP 8-12 error summaries
video demod OP 8-8 error vector, example OP 8-12 error-vector magnitude
See demodulation, digital disk drive
(in symbol table) HT error-vector magnitude, displaying EVM trace HT EXT ARM connector HT EXT REF IN connector HT EXT REF OUT connector HT EXT TRIGGER connector HT external
setting up GS 6-2 disk drives HT about disk drives HT comparison HT copying files HT deleting files HT displaying contents of HT external disk drive HT formatting HT GPIB address: external disk drive HT internal disk drive HT non-volatile RAM disk HT packing a disk HT renaming files HT selecting a default disk HT using disk specifiers HT volatile RAM disk HT volatile RAM disk, removing HT display HT adjusting the color HT allocating portion for programs HT blanking HT changing colors HT cleaning GS 7-19 connecting external GS 7-16 indicators HT See also traces display group hardkeys HT display points, defined HT display resolution OP 14-11, OP 14-12 display state definitions OP 8-10 display, digital OP 14-9 documentation HT DQPSK softkey HT DVB QAM HT block diagram OP 18-4 dynamic signal analyzers OP 13-6
keyboard GS 7-17 monitor GS 7-16 external arm HT about external arm HT trigger, using with external arm HT turning on/off HT external disk
See disk drive EXTERNAL MONITOR connector HT external receiver
See receiver external setup softkey HT external softkey HT external trigger, overview HT eye diagram, example OP 6-5, OP 7-12 eye diagram, using HT
power up GS 7-21 stop frequency is 10 MHz GS 7-23 fault log, displaying HT FFT OP 13-6 properties of OP 13-7 See also spectrum analyzers FFT, overview & terminology HT FILLING TIME RECORD HT filters
digital demodulation OP 17-16, HT gaussian OP 17-18 IF OP 14-18 raised cosine OP 17-17 square-root raised cosine OP 17-17 user defined OP 9-6
GS = Getting Started Guide HT = Online Help OP = Operator's Guide (press (Help) key)
Index

user defined OP 17-18 video OP 14-3 firmware version, displaying HT FM demodulation
carrier offset (in symbol table) HT deviation (in symbol table) HT FSK error (in symbol table) HT FSK error trace, displaying HT FSK measured HT FSK reference HT magnitude error (in symbol table) HT setting the span to symbol rate ratio HT state definitions HT See also demodulation, digital FTP (File Transfer Protocol) HT FTP, and LAN OP 10-10 full span softkey HT fuse GS 7-12
algorithm OP 15-12 example OP 1-6, OP 3-4 FM demodulation, using HT fonts HT formatting a disk GS 6-6 fourier transform OP 13-7 French softkey HT freq counter softkey HT frequency HT about frequency parameters HT center frequency HT frequency counter HT frequency span HT manual frequency, setting HT reference GS 7-7 signal tracking HT span, extending to 8 mhz HT start frequency HT step size HT stop frequency HT stop frequency is 10 MHz GS 7-23 using markers to set step size HT frequency domain overview OP 13-2 frequency error (in symbol table) HT frequency points OP 14-5, HT about frequency points HT and display OP 14-12 memory allocation HT setting the number of HT frequency resolution OP 14-11 frequency response measurements HT frequency response, measuring GS 5-1 frequency span
displaying HT gate delay HT gate time HT gate-delay step size HT setting the length of HT turning on/off HT GATHERING DATA, definition of HT gating
description OP 16-2 example OP 3-3, OP 4-2, GS 3-1 in analog demodulation OP 15-13 gaussian filters OP 17-18 German softkey HT ghosted softkeys HT grids HT 1 grid HT 2 grids HT 4 grids HT hiding HT multiple GS 5-7 overlapped GS 5-7 grounding requirements GS 7-3 group delay softkey HT GSM (Group Service Mobile) HT guardbands, analog demodulation
and display OP 14-5 and information bandwidth OP 14-3 and resolution bandwidth OP 14-14,
OP 14-19 video demodulation OP 17-11, OP 18-12 front panel tour HT
GPIB HT about the GPIB HT addressable only HT devices, setting up GS 6-2 GPIB address: analyzer HT GPIB address: external devices HT
about FSK HT block diagram OP 17-5, HT carrier magnitude-error HT
GS = Getting Started Guide HT = Online Help OP = Operator's Guide (press (Help) key)
Index · filters (continued) · FSK (continued)

GPIB address: external disk drive HT GPIB address: external receiver HT GPIB address: plotter HT GPIB address: printer HT GPIB connector GS 7-16 GPIB controllers HT keys that have GPIB commands HT LAN use OP 10-6 system controller HT
connections HT coupling: ac or dc HT dBm calculations HT disabling HT disabling the anti-alias filter HT input impedance HT input-state table HT input hardkey HT input section (ch1 + j*ch2) softkey HT input state GS 6-9 installation GS 7-9 installing options HT instantaneous spectrum, displaying HT instrument mode hardkey HT interface
H half LED HT description of HT using to set the range HT hardkeys, defined HT Help hardkey HT help, online GS 1-1 hold scale softkey HT horizontal axis
GPIB GS 7-16 LAN GS 7-14 parallel GS 7-15 serial 1 GS 7-15 serial 2 GS 7-7 IQ measured signal OP 17-9, OP 18-10 IQ reference signal OP 17-9, OP 18-10 Italian softkey HT
See X-axis HP-GL (Hewlett-Packard
K keyboard HT
See BASIC I/Q information
about the keyboard HT connecting GS 7-17 keyboard connector HT keyboard type HT knob HT about the knob HT changing the operation of HT
See demodulation, digital ideal state softkey HT IF bandwidth, external receiver HT IF center frequency, external
receiver HT IF section
L LAN GS 7-14, HT about the advanced LAN (UG7) HT about the lan HT activating OP 10-5 addressing OP 10-4 and FTP OP 10-10 connectors, descriptions HT controlling the analyzer via the lan HT ethernet address, displaying HT FTP (File Transfer Protocol), using HT gateway address, setting HT GPIB with OP 10-6 interface, connecting OP 10-1 IP address, setting HT port, configuring HT ports, activating HT LAN (continued)
connecting to RF section GS 7-7 fuse GS 7-12 model number GS 7-10 IF section (0-10 mhz) softkey HT IF trigger
See trigger incoming inspection GS 7-5 indent softkey HT indicators, display HT INF, meaning of HT information bandwidth OP 14-3, OP 14-10 input channels HT about input channels HT channel 2 with rf section HT clearing input overloads HT compensating for gain & delay HT input channels (continued)
GS = Getting Started Guide HT = Online Help OP = Operator's Guide (press (Help) key)
Index · graphics language) HT

ports, selecting HT subnet mask, setting HT Telnet, about HT X-Windows with OP 10-7 X11 display, configuring HT X11 display, control HT X11 display, update rate HT X11 IP address, setting HT leakage OP 14-6 learning products (manuals) HT LEDs: front panel HT level softkey HT line voltage
about markers HT absolute GS 5-4 band power GS 2-3, GS 2-6, HT band power, rms square-root of HT carrier to noise GS 2-4 carrier-to-noise HT carrier-to-noise-density HT coupled, example GS 4-3 frequency counter HT marker into center frequency HT marker into reference level HT marker into span HT marker into start frequency HT marker into stop frequency HT multiple traces HT offset GS 5-5 offset marker, zeroing HT offset markers, repositioning HT offset markers, using HT peak tracking HT polar display markers HT power ratio GS 2-4 relative GS 5-5 repositioning markers HT search GS 5-6 search functions HT trace annotation HT turning on markers HT using for numeric entry HT using to set frequency step size HT using to update frequency values HT waterfall &spectrogram displays HT X-axis, example OP 8-7 x-axis, scaling with markers HT markers, polar
required GS 7-2 RF selector switch GS 7-11 linear averaging
See time averaging lines
See frequency points lines, defined HT listen LED HT LO feedthrough
See zero response loading files
See recalling Local Area Network
See LAN local oscillator OP 13-12
See demodulation, digital magnitude log(dB)/linear softkey HT main length softkey HT main time vs. gate time HT main time-record
See polar markers math HT about math HT cross-correlation math function HT defining a math constant HT displaying math functions HT displaying results HT math constants, overview HT math function operands HT math functions HT memory for math-buffers HT recalling math functions HT saving math functions HT user defined GS 6-7 waterfall/spectrogram displays HT MAU connector HT
displaying HT length equal to gate length HT main length HT main time HT setting the length of HT See also time record manual sweep OP 14-13 manual sweep, selecting HT manuals, for this product HT Marker Cnt, meaning of HT marker group hardkeys HT Marker Man, meaning of HT marker readout HT Marker/Entry LED HT markers HT
GS = Getting Started Guide HT = Online Help OP = Operator's Guide (press (Help) key)
Index

measured filter, selecting HT measured signal
See LAN NO DATA, definition of HT noise equivalent bandwidth OP 14-18 noise, measuring OP 4-1, GS 2-1 non-volatile RAM disk
video demod OP 18-10 measured signal, digital demod OP 17-9 measurement calculations, disabling
HT measurement data, about HT MEASUREMENT Group hardkeys HT measurement points, defined HT measurement resolution OP 14-11 measurement speed, enhancing HT measurement state GS 6-9, HT recalling HT saving HT measurement time OP 14-2, OP 14-10, OP 14-17 measurements HT continuing HT displaying meas-state table HT pausing HT single-stepping HT starting HT memory
See disk drives normalization OP 8-4, HT numeric entry softkeys HT
video demod OP 7-13 offset marker
See markers offset, in digital demod OP 6-8 online help GS 1-1 See Help hardkey operating environment GS 7-9, GS 7-19 options, installing HT options, temporary HT origin offset
See demodulation, digital output filter on/off softkey HT output z softkey (source) HT OV1 or OV2, definition of HT oven frequency reference GS 7-7 OVEN REF OUT connector HT over LED HT description of HT using to set the range HT overlap processing HT
about memory HT capture RAM HT erasing HT for applications HT for frequency points HT for temporary math buffers HT memory limitations OP 14-19 menu HT minimum loss pad GS 7-18 mirror freq on/off softkey HT mirrored spectrums OP 18-17 Mkr Val HT model number
P packing a disk HT parallel devices, setting up GS 6-2 parallel port GS 7-15, HT pausing a measurement HT PCL (HP-GL) HT PCL (printer control language) HT PDC (Personal Digital Cellular) HT peak search HT peak track softkey HT peak tracking, example OP 3-3 peak-hold averaging
IF section GS 7-10 RF section GS 7-11 monitor
connecting external GS 7-16 MSK softkey HT
Cellular) HT NADC demodulation, example OP 6-4 NAN, meaning of HT narrowband measurements OP 13-8 network measurements GS 5-1
See averaging performance test softkey HT periodic chirp softkey HT peripherals, use of GS 6-2
phase GS 5-7 displayed GS 4-2
GS = Getting Started Guide HT = Online Help OP = Operator's Guide (press (Help) key)
Index · Network, Local Area

relative, example GS 4-1 wrap GS 5-7 phase error
power ratio markers GS 2-4 power spectral density GS 2-2 pre-trigger delay HT preset hardkey HT printer
See demodulation, digital phase noise, analysis example OP 2-1 phase, displaying HT PHS (PHP) HT plotter
interface GS 7-15 setting up GS 6-2 printing GS 6-3, HT aborting HT about printing HT data and file formats HT date and time HT online help, printing HT output to file HT selecting an output device HT selecting display items HT setting the GPIB address HT starting HT PROBE POWER connectors HT problems, digital/video demod HT PSD measurements HT psk (phase shift keying) HT pulse length, in digital demod OP 6-6 pulse modulation
setting up GS 6-2 plotter interface GS 7-15 plotting GS 6-3, HT aborting HT about plotting HT changing plotter pen assignments HT data and file formats HT date and time HT default pen assignments HT output to file HT plot speed HT selecting an output device HT selecting display items HT setting P1/P2 HT setting the GPIB address HT specifying line types HT starting HT PM demodulation
See demodulation, digital PULSE NOT FOUND HT pulse search
algorithm OP 15-10 example OP 1-4, OP 2-3 using HT points
in digital demod OP 17-14 setup example OP 6-6 pulsed signals, digital demod OP 17-14
See frequency points points, defined HT points-per-symbol, setting OP 8-4, HT polar display, using HT polar markers
block diagram, video demod OP 18-4 QAM HT QAM demodulation, example OP 8-1 QPSK softkey HT
example OP 8-4 units, example OP 8-4 post-trigger delay HT power HT
adjacent-channel GS 2-6 band power GS 2-6, HT band power, rms square-root of HT carrier-to-noise HT carrier-to-noise-density HT consumption GS 7-2 cords GS 7-3 power ratio GS 2-6, HT turn on failure GS 7-21
See memory RAM disk
See disk drives random noise softkey HT range HT autoranging HT example GS 5-3 setting optimum range HT single ranging HT range (continued)
tracking the reference level HT
GS = Getting Started Guide HT = Online Help OP = Operator's Guide (press (Help) key)
Index · power calculation, digital · demodulation OP 8-4

range hardkey HT RBW
and gating OP 16-5 coupling to frequency span HT digital demod OP 17-11 effect on noise OP 14-2 entering arbitrary values HT interaction with other parameters HT scalar limitations OP 14-11 setting the resolution bandwidth HT vector limitations OP 14-18 video demod OP 18-11 See also bandwidth coupling result length
See resolution bandwidth real-time bandwidth HT rear panel tour HT recalling HT about recalling HT autostate file HT BASIC programs HT capture-buffer contents HT math functions HT measurement state HT recalling data GS 6-5 spectrogram displays OP 5-16 trace GS 3-2, HT waterfall displays OP 5-16 receiver HT
digital demod OP 6-8, OP 17-14 video demod OP 7-13 result length softkey HT RF section
connecting to IF section GS 7-7 fuse GS 7-13 model number GS 7-11 voltage selector switch GS 7-11 rms averaging
about receivers HT GPIB address: external receiver HT using an external receiver HT receiver softkey HT REF, definition of HT reference filter, selecting HT reference level
See averaging root raised cosine filters OP 17-17 RPG
See reference line reference line HT displaying HT examples of HT position of HT setting reference level HT setting with the marker HT tracking the range HT reference signal
See inside front cover sample frequency
See sample rate sample rate OP 14-14 video demodulation OP 17-11, OP 18-12 samples, defined HT saving HT about saving HT autostate file HT calibration trace HT capture buffer contents HT math functions HT measurement state HT re-saving BASIC programs HT saving BASIC programs HT saving data GS 6-4 traces HT
digital demod OP 17-9 video demod OP 18-10 reference, frequency GS 7-7 references
digital demod OP 17-2 Remote LED HT REMOTE OPERATION group hardkeys HT requirements
grounding GS 7-3 power GS 7-2 resolution
display OP 14-12 frequency OP 14-11 measurement OP 14-5
scalar measurements HT about scalar mode HT
resolution bandwidth HT about resolution bandwidth HT
GS = Getting Started Guide HT = Online Help OP = Operator's Guide (press (Help) key)
Index · See knob RS232 · See serial port

block diagram HT block diagrams OP 12-3 limitations OP 14-11 overview OP 12-2 scale at markers softkey HT scaling HT auto scaling HT x-axis HT x-axis scaling: linear vs. log HT X-axis, example OP 8-7 y-axis scaling HT scan time
am modulating the output HT amplitude, setting HT arbitrary waveforms HT arbitrary waveforms, duration HT available source outputs HT connections HT CW (fixed sine) output HT dc offset, setting HT displaying the source-state table HT external signals, using as input HT output filter, disabling HT output impedance & dBm units HT output impedance, setting HT output protection, clearing HT periodic chirp & frequency span HT periodic chirp output HT periodic chirp, duration HT random noise output HT sine frequency, setting HT SOURCE LED HT turning on and off HT source hardkey HT source type softkey HT source, setup example GS 5-3 span HT arbitrary spans HT cardinal spans HT coupling to main length HT in analog demodulation OP 15-6 maximum span HT setting span HT setting with the marker HT video demodulation OP 17-11, OP 18-12 See also frequency span span softkey HT span, 8 MHz (opt. AYH) OP 18-18 Spanish softkey HT spectral displays OP 5-1 spectral map
See sweep time screen, cleaning GS 7-19 search length HT
digital demod OP 6-8 video demod OP 7-13 search length, in digital demod
OP 17-14 self test HT long confidence test HT quick confidence test HT self-test log HT serial 1 port GS 7-15 serial 1/serial 2 connectors HT serial 2 port GS 7-7 serial devices, setting up GS 6-2 serial number, displaying HT serial number, location HT serial port, configuring & cabling HT serial x setup softkey HT shifted functions HT shipping GS 7-20 signal to noise markers GS 2-4 signal track softkey HT signal tracking HT sine freq softkey HT single measurements, running HT single range softkeys HT single ranging HT single sweep, selecting HT SNR, digital/video demodulation HT softkeys HT about softkeys HT bracketed softkeys HT numeric entry HT softkeys that toggle HT
See waterfall spectrogram HT about spectrogram displays HT colorbar HT colors, selecting HT colors, setting the number of HT displaying OP 5-1 enhancing HT markers, using HT spectrogram (continued)
source HT about the source HT
number of traces, setting HT
GS = Getting Started Guide HT = Online Help OP = Operator's Guide (press (Help) key)
Index

pausing HT printing/plotting, formats HT printing/plotting, overview HT real-time indicator HT recalling HT saving HT threshold, setting HT trace buffers, removing HT trace, selecting HT turning on HT using in math functions HT spectrum analyzers
symbol states, example OP 8-8 symbol table
See demodulation, digital symbol/error table, example OP 8-8 SYNC NOT FOUND HT SYNC NOT FOUND, pulse search HT SYNC OUT connector HT sync pattern
digital demod, OP 6-9 video demod, OP 7-14 See also demodulation, digital See also demodulation, video sync search
comparisons OP 13-9 types of OP 13-4 spectrum measurements HT speed, enhancing HT square-root raised cosine filters
example OP 6-8, OP 7-13 in digital demod OP 17-13 video demod OP 18-15 See also demodulation, digital See also demodulation, video sync word
OP 17-17 SRQ LED HT standard setups
digital demod OP 6-8 video demod OP 7-13 SYSTEM Group hardkeys HT system interconnect GS 7-16 SYSTEM INTERCONNECT connector
digital demodulation, example OP 6-4 start frequency HT setting start frequency HT setting with the marker HT starting a measurement HT state definitions HT See display state definitions See input state See measurement state stop frequency HT setting stop frequency HT setting with the marker HT suffix menus HT Swedish softkey HT sweep speed
T talk LED HT TDMA
See also NADC Telnet, about HT testing
See self test ThinLAN connector GS 7-14, HT TIFF (tag-based file format)) HT time averaging OP 14-15 See averaging time capture HT about time capture HT calibration and time-capture HT displaying capture information HT displaying capture-buffer contents HT playback example OP 2-2 recalling capture-buffer contents HT recalling data from disk OP 2-2 saving capture-buffer contents HT starting a time capture HT time capture buffer HT time corrections HT time corrections, analog demodulation
See sweep time sweep time
described OP 14-10 See also measurement time sweep, manual OP 14-13 sweeping HT about sweeping HT auto sweep, selecting HT continuous sweep, selecting HT manual sweep, selecting HT single sweep, selecting HT symbol clock, adjusting HT symbol conventions
See display state definitions symbol rate OP 17-11, OP 18-12 symbol rate softkey HT
GS = Getting Started Guide HT = Online Help OP = Operator's Guide (press (Help) key)
Index

time data OP 14-15 time data softkey HT time domain overview OP 13-2 time gating HT about time gating HT displaying gate time-record HT gate delay HT gate length HT selecting a window HT turning on/off HT See also gating time points, digital & video demodulation HT time record OP 13-7, OP 14-16, HT
disabling meas. calculations HT displaying group delay HT displaying linear magnitude HT displaying log magnitude HT displaying multiple traces HT frequency response HT hiding HT imaginary part displaying HT instantaneous spectrum HT marker annotation HT math functions HT modifying trace labels HT multiple GS 5-7 multiple active GS 5-7 overlaid OP 6-10 overlaying HT phase, displaying HT PSD HT real part (real data), dispaying HT recalling HT saving HT scaling data, overview HT selecting measurement data HT spectrum HT time-domain data HT trace boxes HT See also display See also grids transporting GS 7-20 trellis diagram, using HT trigger HT
about the time record HT applying a window HT coupling to span HT displaying gate time-record HT displaying main time-record HT effects of averaging HT effects of demodulation HT effects of span HT FILLING TIME RECORD message HT gate length HT main length HT real vs. complex data HT time record length
and memory OP 14-19 defined OP 14-16 limitations OP 14-19 time record size
defined OP 14-16 time resltn softkey HT time softkey (averaging) HT time, changing HT time-domain analysis HT time-domain data, displaying HT traces HT about traces HT activating multiple traces HT adding a title HT adding trace information HT auto correlation HT capture buffer contents HT changing colors HT coherence HT coordinates HT cross correlation HT cross spectrum HT traces (continued)
about trigger HT analog demodulation OP 15-13, HT available trigger signals HT delay HT EXT TRIGGER LED HT external arm, using with trigger HT external trigger HT free run trigger HT GPIB trigger HT IF channel trigger HT input channel trigger HT level HT slope HT source trigger HT trigger holdoff HT trigger LED HT trigger line HT trigger type softkey HT troubleshooting, digital demod HT two-channels
data register contents HT
GS = Getting Started Guide HT = Online Help OP = Operator's Guide (press (Help) key)
Index

digital demod OP 6-12 video demod OP 7-15
saving HT setting a threshold HT skewing HT spacing of traces HT trace buffers, removing HT trace height HT trace, selecting HT traces, position and baseline HT using in math functions HT when you can use them HT y-axis scaling HT window bandwidth
U U.K. English softkey HT U.S. English softkey HT UFG (LAN and 4 MB memory option)
HT UG7 (advanced LAN option) HT UNCAL, definition of HT units: x-axis HT units: y-axis HT user defined filters OP 17-18
defined OP 14-18 values OP 14-18 windows HT about windows HT comparison HT equivalent noise bandwidth HT example GS 5-3 window bandwidth HT window functions OP 14-6 window shapefactor HT windows for gate time record HT windows for main time record HT
See time averaging vector diagram, example OP 6-5, OP 7-12 vector measurements OP 13-8, HT about vector mode HT block diagram HT block diagrams OP 12-5 overview OP 12-4 video averaging OP 14-3 video demodulation
See demodulation, video video filtering OP 14-3 view state GS 6-9 volatile RAM disk
See scaling X-axis, scaling and markers OP 8-7 x-axis, scaling with markers HT X-Windows, LAN use OP 10-7 X11 display, using
See disk drives voltage selector switch
RF section GS 7-11 VSB
block diagram OP 18-6 center frequency, setting OP 7-4 VSB 8/16 softkey HT
W waterfall HT about waterfall displays HT baselines, showing HT displaying OP 5-1 elevation HT hiding traces HT markers, using HT number of traces, setting HT pausing HT printing/plotting, formats HT printing/plotting, overview HT waterfall (continued)
Z zero padding OP 14-20 zero response OP 14-9 zero span measurements OP 15-2, HT zoom measurements OP 13-8, OP 14-3, HT
GS = Getting Started Guide HT = Online Help OP = Operator's Guide (press (Help) key)



If you are thinking about... And you want to... Then read the analyzer’s...
Install the analyzer, or do operation verification or performance verification tests
Installation and Verification Guide
Getting started Make your first measurements with your new analyzer
Review measurement concepts Operator’s Guide
Learn what each key does Online Help (press the [ Help ] key)
Making measurements Learn how to make typical measurements
Getting Started Guide and Operator’s Guide
Learn the Agilent Instrument BASIC interface
Agilent 89400-Series Using Agilent Instrument BASIC
(To receive Agilent Instrument BASIC
Program with Agilent Instrument BASIC
Agilent Instrument BASIC User’s Handbook
and Agilent Instrument BASIC manuals, order option 1C2)
Remote operation Learn about the GPIB and SCPI GPIB Programmer’s Guide
Find specific GPIB commands quickly Agilent 89400-Series GPIB Commands: Quick Reference
Find GPIB command details Agilent 89400-Series GPIB Command Reference
Transfer analyzer data to or from a PC (Personal Computer) application
Using analyzer data with a PC
Display analyzer data on a PC, or display PC data on the analyzer
Adjust, troubleshoot, or repair the analyzer
Servicing the analyzer (To receive service information,

If you need assistance, contact your nearest Agilent Technologies Sales and Service Office listed in the Agilent Catalog. You can also find a list of local service representatives on the Web at: http://www.agilent-tech.com/services/English/index.html or contact your nearest regional office listed below.
If you are contacting Agilent Technologies about a problem with your Agilent 89410A Vector Signal Analyzer, please provide the following information: Model number: Agilent 89410A Serial number: Options: Date the problem was first encountered: Circumstances in which the problem was encountered: Can you reproduce the problem? What effect does this problem have on you?
You may find the serial number and options from the front panel of your analyzer by executing the following:
Press [System Utility], [more], [serial number].
Press [System Utility], [options setup].
If you do not have access to the Internet, one of these centers can direct you to your nearest representative:
United States Test and Measurement Call Center (800) 452-4844 (Toll free in US)
Europe (31 20) 547 9900
Japan Measurement Assistance Center (81) 426 56 7832 (81) 426 56 7840 (FAX)
Latin America (305) 267 4245 (305) 267 4288 (FAX)
Australia/New Zealand 1 800 629 485 (Australia) 0800 738 378 (New Zealand)
Asia-Pacific (852) 2599 7777 (FAX) (852) 2506 9285
Canada (905) 206-4725