Agilent 8590 E-Series Portable Spectrum Analyzers Data Sheet

Agilent Technologies 8595E — Agilent 8590 E-Series Portable Spectrum Analyzers Data Sheet diagram (page 1)
Fig. 2 · Agilent 8590 E-Series Portable Spectrum Analyzers Da…

These specifications apply to the Agilent Technologies 8591E, 8593E, 8594E, 8595E, and 8596E spectrum analyzers.

8594E dc coupled 9 kHz to 2.9 GHz ac coupled 100 kHz to 2.9 GHz 8595E dc coupled 9 kHz to 6.5 GHz ac coupled 100 kHz to 6.5 GHz 8596E dc coupled 9 kHz to 12.8 GHz ac coupled 100 kHz to 12.8 GHz Band LO harmonic = N 0 1 9 kHz to 2.9 GHz (dc coupled) 0 1 100 kHz to 2.9 GHz (ac coupled) 1 1 2.75 GHz to 6.5 GHz 2 2 6.0 GHz to 12.8 GHz

Specifications All specifications apply over O°C to +55°C. The analyzer will meet its specifications after 2 hours of storage at a constant temperature, within the operating temperature range, 30 minutes after the analyzer is tumed on, and after CAL FREQ and CAL AMPTD (and for the 8593E, 8595E, and 8596E CAL YTF) have been run.

Frequency Specifications Frequency Range 8591E 50 Ω 9 kHz to 1.8 GHz 75 Ω 1 MHz to 1.8 GHz 8593E 9 kHz to 22 GHz Option 026/027 9 kHz to 26.5 GHz

Frequency Reference (Opt. 004) Aging ±2 x 10-6/year ±1 x 10-7/year Temperature Stability ±5 x 10-6 ±1 x 10-8

Initial Achievable Accuracy ±0.5 x 10-6 ±2.2 x 10-8

Agilent Technologies 8595E — Agilent 8590 E-Series Portable Spectrum Analyzers Data Sheet diagram (page 1)
Fig. 3 · Agilent 8590 E-Series Portable Spectrum Analyzers Da…

Band LO harmonic = N 0 1 9 kHz to 2.9 GHz 1 1 2.75 GHz to 6.5 GHz 2 2 6.0 GHz to 12.8 GHz 3 3 12.4 GHz to 19.4 GHz 4 4 19.1 GHz to 22.0 GHz 4 4 (Opt. 026/027) 19.1 GHz to 26.5 GHz

Page 2 of the Agilent Technologies 8595E user manual — Frequency Readout
Page 2 · click to enlarge
Page 2

Frequency Readout

Accuracy ±20% Selectivity (Characteristic)

Accuracy (Start, Stop, Center, Marker) ±(frequency readout x frequency reference error1+span accuracy +1% of span +20% of RBW+100 Hz x N*)

3 kHz to 10 kHz 15:1 100 kHz to 3 MHz 15.1 1 kHz, 30 kHz 16:1 –40 dB/–3 dB

Marker Count Accuracy Frequency Span

30 Hz to 300 Hz 10:1

≤10 MHz x N* ±(marker frequency x frequency reference error1+ counter resolution +100 Hz x N*) Frequency Span ±(marker frequency x frequency >10 MHz x N* reference error1+ counter resolution +1 kHz x N*) Counter Resolution

Video Bandwidth Range 30 Hz to 1 MHz in 1,3 sequence 1 Hz to 1 MHz (Opt 130) Stability Noise Sidebands (1 kHz RBW, 30 Hz VBW and sample detector) >10 kHz offset from CW signal ≤–90 dBc/Hz + 20 Log N* >20 kHz offset from CW signal ≤–100 dBc/Hz + 20 Log N* >30 kHz offset from CW signal

Frequency Span ≤10 MHz x N* Selectable from 10 Hz to 100 kHz Frequency Span >10 MHz x N* Selectable from 100 Hz to 100 kHz

Agilent Technologies 8595E — Frequency Readout diagram (page 2)
Fig. 4 · Frequency Readout

Frequency Span Range 0 Hz (zero span), and

Opt. 130 Std. Min. Min. Max (KHz) (KHz) (GHz)

8591E 1 10 1.8 8593E 1 x N* 10 x N* 19.25 8594E 1 10 2.9 8595E 1 10 6.5 8596E 1 x N* 10 x N* 12.8

Resolution Four digits or 20 Hz x N* whichever is greater Accuracy

8591E 1 kHz RBW, 1 kHz VBW ≤250 Hz pk-pk in 100 ms 30 Hz RBW, 30 Hz VBW ≤30 Hz pk-pk in 300 ms 8593E, 94E, 95E, 96E 1 kHz RBW, 1 kHz VBW ≤(250 x N*) Hz pk-pk in 100 ms 30 Hz RBW, 30 Hz VBW ≤(30 x N*) Hz pk-pk in 300 ms System-Related Sidebands

Span ≤10 MHz x N* ±2% of span Span >10 MHz x N* ±3% of span

Frequency Sweep Time Range

Span = 0 Hz, >1 kHz 20 ms to 100 s Span = 0 Hz (Opt. 101) 20 µs to 100 s Accuracy

>30 kHz offset from CW signal ≤–65 dBc + 20 Log N*

20 ms to 100 s ±3% 20 µs to

Comb Generator Frequency 8593E, 96E I00 MHz fundamental frequency Accuracy ±0.007%

* N = LO harmonic. N = 1 for 91E, 94E, 95E 1. Frequency reference error = (aging rate x period of time since adjustment + initial achievable accuracy + temperature stability).

–60 dB/–3 dB · Residual FM

Page 3 of the Agilent Technologies 8595E user manual — Mixer Level (dBm)
Page 3 · click to enlarge
Page 3

Amplitude Specifications Amplitude specifications do not apply for Analog+ mode and negative peak detector mode except as noted in “Amplitude Characteristics.”

Displayed Average Noise Level (Input terminated, 0 dB attenuation, 1 Hz/30 Hz VBW, sample- detector) 8591E 30 Hz RBW 1 kHz RBW 400 kHz to 1 MHz ≤–130 dBm ≤–115 dBm 1 MHz to 1.5 GHz ≤–130 dBm ≤–115 dBm 1.5 GHz to 1.8 GHz ≤–128 dBm ≤–113 dBm 8591E (Opt. 001) 1 MHz to 1.5 GHz ≤–78 dBmV ≤–63 dBmV 1.5 GHz to 1.8 GHz ≤–76 dBmV ≤–61 dBmV 8593E 400 kHz to 2.9 GHz ≤–127 dBm ≤–112 dBm 2.75 GHz to 6.5 GHz ≤–129 dBm ≤–114 dBm 6.0 GHz to 12.8 GHz ≤–117 dBm ≤–102 dBm 12.4 GHz to 19.4 GHz ≤–113 dBm ≤–98 dBm 19.1 GHz to 22 GHz ≤–107 dBm ≤–92 dBm 8593E (Opt. 026/027) 19.1 GHz to 26.5 GHz ≤–102 dBm ≤–87 dBm 8594E 400 kHz to

Amplitude Range

Displayed average noise level to +30 dBm 8591E (Opt. 001) Displayed average noise level to +72 dBmV Maximum Safe Input Level (input attenuator ≥10 dB) Average Continuous Power +30 dBm (1 W) 8591E (Opt. 001) +72 dBmV (0.2 W) Peak Pulse Power

8591E +30 dBm (1 W) 8591E (Opt. 001) +72 dBmV (0.2 W) 8593E, 94E, 95E, 96E +50 dBm (100 W) for < 10 µs pulse width and

8591E 25 Vdc 8591E (Opt. 001) 100 Vdc 8593E 0 Vdc 8594E, 95E, 96E 0 V (dc coupled) 50 V (ac coupled)

Gain Compression >10 MHz ≤0.5 dB (total power at input mixer2 = –10 dBm)

Nominal Dynamic Range

Agilent 8593E (Option 130)

Sensitivity 30 Hz RBW Third Order Intermod Second Order Distortion

10 MHz to 2.9 GHz

Dynamic Range (dB)

(30 Hz Res BW)

19.1 TO 22 GHz

12.4 TO 19.4 GHz

>2.75 GHz

6.0 TO 12.8 GHz

400 kHz TO 6.5 GHz

–70 –60 –50 –40 –30 –20 –10 0 +10

Mixer Level (dBm)

2. Mixer Power Level (dBm) = Input Power (dBm) Input Atten. (dB)
Page 4 of the Agilent Technologies 8595E user manual — Amplitude Range
Page 4 · click to enlarge
Page 4

Spurious Responses Second Harmonic Distortion

Reference Level Range same as amplitude range Resolution 0.1 dB for log scale, 0.12% of reference level for linear scale Accuracy ±0.3 dB @ –20 dBm 0 dBm to –59.9 dBm ±(0.3 dB +.01 x dB from –20 dBm)

5 MHz to 1.8 GHz (91E)

10 MHz to 2.9 GHz (93E) 10 MHz (94E, 95E, 96E) input mixer.2

>2.75 GHz (93E, 95E,96E)

Frequency Response (10 dB input attenuation) 8591E Absolute3 Relative Flatness4

9 kHz to 1.8 GHz ±1.5 dB ±1.0 dB 8593E Preselector peaked in band > 0 Absolute3 Relative Flatness4

5 MHz to 1.8 GHz (91E) 10 MHz (93E, 94E, at input mixer2 and >50 kHz 95E, 96E) separation. Other Input Related Spurious

9 kHz to 2.9 GHz ±1.5 dB ±1.0 dB 2.75 GHz to 6.5 GHz ±2.0 dB ±1.5 dB 6.0 GHz to 12.8 GHz ±2.5 dB ±2.0 dB 12.4 GHz to 19.4 GHz ±3.0 dB ±2.0 dB 19.1 GHz to 22 GHz ±3.0 dB ±2.0 dB 19.1 GHz to 26.5 GHz ±5.0 dB ±2.0 dB 8594E, 95E, 96E (dc coupled preselector peaked) Absolute3 Relative Flatness4

≤1.8 GHz (91E)

≤6.5 GHz (95E) ≤12.8 GHz (96E) ≤18 GHz (93E) ≤22 GHz (93E)

9 kHz to 2.9 GHz ±1.5 dB ±1.0 dB 2.75 GHz to 6.5 GHz ±2.0 dB ±1.5 dB 6.0 GHz to 12.8 GHz ±2.5 dB ±2.0 dB

Residual Responses (input terminated and 0 dB attenuation) 1 MHz to 1.8 GHz (91E Opt. 001)

Calibrator Output Amplitude –20 dBm ±0.4 dB 8591E Opt.001 +28.75 dBmV ±0.4 dB

Resolution Bandwidth Switching Uncertainty (Referenced to 3 kHz RBW, at ref. level) 3 kHz to 3 MHz RBW ±0.4 dB 1 kHz RBW ±0.5 dB 30 Hz to 300 Hz RBW ±0.6 dB

Display Range Log Scale 0 to –70 dB from reference level is calibrated; 0.1, 0.2, 0.5 dB/division and 1 to 20 dB/division in 1 dB steps; eight divisions displayed. Linear Scale Eight divisions Scale units dBm, dBmV, dBuV, V, and W

Linear to Log Switching ±0.25 dB at reference level

Display Scale Fidelity Log Maximum Cumulative

Marker Readout Resolution 0.05 dB for log scale 0.05% of reference level for linear scale Fast Sweep Times for Zero Span (Opt. 101 or 301)

0 to –70 dB from reference level 3 kHz to 3 MHz RBW ± (0.3 + 0.01 x dB from reference level) 30 Hz to 1kHz RBW ± (0.4 + 0.01 x dB from reference level) Log Incremental Accuracy ±0.4 dB/4 dB 0 to –60 dB from reference level Linear Accuracy ±3% of reference level

20 µs to 20 ms ≤1 GHz 0.7% of reference level for linear scale >1 GHz 1.0% of reference level for linear scale

3. Referenced to 300 MHz CAL OUT. 4. Ref. to midpoint between highest and lowest freq. response deviations.
Page 5 of the Agilent Technologies 8595E user manual — General Specifications Temperature Range
Page 5 · click to enlarge
Page 5

Option Specifications Option 010 and 011 Tracking Generator Frequency Range 8591E 100 kHz to 1.8 GHz (Opt. 011, 75 Ω) 1 MHz to 1.8 GHz 8593E, 94E, 95E, 96E 9 kHz to 2.9 GHz

Dynamic Range5 TG Feedthrough 8591E 106 dB ≤–106 dBm 8591E (Opt. 011) 100 dB ≤–57.24 dBmV 8593E (> 400 kHz) 111 dB ≤–112 dBm 8594E (> 400 kHz) 106 dB ≤–107 dBm (> 5 MHz) 111 dB ≤–112 dBm 8595E (>400 kHz) 109 dB ≤–110 dBm 8596E (> 400 kHz) 109 dB ≤–110 dBm

8591E 0 to –70 dBm 8591E (Opt. 011) +42.8 to –27.2 dBmV 8593E, 94E, 95E, 96E –1 to –66 dBm Resolution 0.1 dB Absolute Accuracy (@ 300 MHz, –20 dBm, +28.8 dBmV)

8591E (–15 dBm to 0 dBm) –(source attenuator setting) 8591E (Opt 011) (+27.8 to 42.8 dBmV)–(source attenuator setting) 8593E, 94E, 95E, 96E (–10 dBm to –1 dBm)–(source attenuator setting) Resolution 0.1 dB

8591E ±1.0 dB 8593E, 94E, 95E, 96E ±0.75 dB

8591E 10 dB 8593E, 94E, 95E, 96E 9 dB Accuracy

Option 103 Quasi-Peak Detector

Amplitude response conforms with Publication 16 of Comité International Spécial des Perturbations Radioélectriques (CISPR) Section 1, Clause 2.

8591E ±0.75 dB 8593E, 94E, 95E, 96E ±0.5 dB

Option 105 Time Gated Spectrum Analysis Gate Delay

Range 1 µs to 65.535 ms Resolution 1 µs Accuracy ±(1 µs + 0.01% x Gate Delay Readout)6

8591E 0 to 60 dB, 10 dB steps 8593E, 94E, 95E, 96E 0 to 56 dB, 8 dB steps

Output Flatness

(From Gate Trigger Input to positive edge of Gate Output)

8591E ±1.75 dB 8593E, 94E, 95E, 96E (>10 MHz) ±2.0 dB

Gate Length

Range 1 µs to 65.535 ms Resolution 1 µs Accuracy ±(0.2 µs + (0.01% x Gate Length Readout)) (From positive edge to negative edge of Gate Output)

Effective Source Match (Characteristic)

8591E 1.6:1 (10 dB attenuation) 8593E, 94E, 95E, 96E 1.5:1 (8 dB attenuation)

Spurious Output Harmonic Spurs

Additional Gate Amplitude Error 7

Log Scale

8591E (0 dBm, +42.8 dBmV output)

General Specifications Temperature Range

8591E

Operating 0°C to +55°C Storage –40°C to +75°C

EMI Compatibility Conducted and radiated interference CISPR Pub. 11 and Messempfaenger Postverfuegung 526/527/79.

Audible Noise

5. Maximum output level minus TG feedthrough. 6. Up to 1 Vs jitter due to 1 µs resolution of gate delay clock. 7. With GATE ON enabled and triggered, CW Signal, Peak Detector Mode.
Show 5 more labels / short text on this page
Dynamic Range (Characteristic) · Output Level Range · Power Sweep Range · Vernier Range · Output Attenuator Range
Page 6 of the Agilent Technologies 8595E user manual — lnputs/Outputs
Page 6 · click to enlarge
Page 6

Power Requirements ON (Line 1) 90 to 132 V rms, 47 to 440 Hz 195 to 250 V rms, 47 to 66 Hz Power consumption

Keyboard (Opt. 041 or 043) 5 Pin mini-DIN, compatible with HP C1405B and most IBM/AT key boards Ext. Trigger Input BNC, TTL levels, positive edge trigger GPIB and Parallel (Opt 041) SH1, AH1, T6, L4, ST1, RL1, PPO, DC1, C1 C2, C3, & C28 and 25 Pin subminature D-shell female for parallel RS-232 and Parallel (Opt 043) 9 Pin subminature D-shell female

User Program Memory 238 Kbytes non-volatile RAM

Internal 24 traces or 32 states External 50 traces, 8 states Memory card (85700A) 32 Kbytes

and 25 Pin subminature D-shell female for parallel Ext Ref In 50 ΩBNC, 10 MHz, –2 to +10 dBm 10 MHz Ref Output 50 ΩBNC, 10 MHz, 0 dBm Aux Interface 9 pin “D” subminiature Pin 1-4, TTL Output Pin 5 TTL Input Pin 6 Gnd Pin 7 –15 vdc ±5%; 150 mA max Pin 8 +5 vdc ±5%; 150 mA max Pin 9 +15 vdc ±5%; 150 mA max Monitor Out 50 ΩBNC, Selectable Format NTSC, 15.75 kHz, 60 Hz PAL, 15.625 kHz, 50 Hz Dimensions (Nominal) (Without handle, feet, 163 mm (H) x 325 mm (W) or cover) x 427 mm (D)

lnputs/Outputs

Front Panel Connectors Input 50 ΩType N (Opt 001) 75 ΩBNC female (Opt 026) APC 3.5 mm male (Opt 027) 50 ΩType N female Cal Output 50 ΩBNC, –20 dBm, 300 MHz 100 MHz Comb Out 100 MHz ±0.007%, SMA Probe Power +15 Vdc, –12.6 Vdc, and Gnd (150mA max each)

Rear Panel Connectors Earphone (Opt 102 and 103) 1/8 inch monaural jack LO Output (Opt 009) 50 ΩSMA Female, 3.0 to 6.8214 GHz TV Trigger Output BNC, TTL levels, negative edge (Opt 101 and 102) trigger after sync pulse Gate Trigger Input (Opt 105) 50 ΩBNC, Pulsewidth >30 ns (TTL) Gate Output (Opt 105) 50 ΩBNC (TTL) SWEEP + Tune Output

(Overall) 184 mm (H) x 373 mm (W) X 461 mm (D) Weight (Nominal) 8591E 15.4 kg (34 lb) 8593E 16.4 kg (36 lb) 8594E 16.4 kg (36 lb) 8595E 16.4 kg (36 lb) 8596E 16.4 kg (36 lb)

(Opt 009) 2 k ΩBNC, 0 to +10V, 0.36V/GHz of CF Ext. ALC Input 1 MW, –66 dBV to +6 dBV Sweep Output BNC, 5 k W, 0 to +10 V ramp High Sweep In/Out BNC, high TTL = sweep, low TTL = Retrace Aux Video Out 50 ΩBNC, 0-1 V Uncalibrated Aux IF Output 50 ΩBNC, –10 to –60 dBm, 21.4 MHz

Agilent Technologies 8595E — lnputs/Outputs diagram (page 6)
Fig. 5 · lnputs/Outputs

Data Storage (nominal)

Page 7 of the Agilent Technologies 8595E user manual — lnputs/Outputs
Page 7 · click to enlarge
Page 7

Related Literature Pub. Number General Purpose Information Agilent 8590L and 8592L Product Overview 5962-7275E Agilent 8590C/E/L and EM Series Configuration Guide 5963-6858E Agilent 8590 E-Series Brochure 5963-6908E

Agilent Technologies’ Test and Measurement Support, Services, and Assistance Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We strive to ensure that you get the test and measurement capabilities you paid for and obtain the support you need. Our extensive support resources and services can help you choose the right Agilent products for your applications and apply them successfully. Every instrument and system we sell has a global warranty. Support is available for at least five years beyond the produc- tion life of the product. Two concepts underlie Agilent’s overall support policy: “Our Promise” and “Your Advantage.”

Product Feature Briefs Analog + Display 5091-4054E Transmitter Power Measurements (ACP, OBW) 5091-4055E Zoom Window 5091-4051E Measuring AM with FFT 5091-4049E Time Gated Spectrum Analysis 5091-4053E Editing Keyboard 5091-4048E Marker and Peaks Table 5091-4050E Third Order Intermodulation, N & B Bandwidth, and Percent AM 5091-4052E

Our Promise “Our Promise” means your Agilent test and measurement equip- ment will meet its advertised performance and functionality. When you are choosing new equipment, we will help you with product information, including realistic performance specifica- tions and practical recommendations from experienced test engineers. When you use Agilent equipment, we can verify that it works properly, help with product operation, and provide basic measurement assistance for the use of specified capabili- ties, at no extra cost upon request. Many self-help tools are available.

Your Advantage “Your Advantage” means that Agilent offers a wide range of additional expert test and measurement services, which you can purchase according to your unique technical and business needs. Solve problems efficiently and gain a competitive edge by contracting with us for calibration, extra-cost upgrades, out- of-warranty repairs, and on-site education and training, as well as design, system integration, project management, and other professional services. Experienced Agilent engineers and tech- nicians worldwide can help you maximize your productivity, optimize the return on investment of your Agilent instruments and systems, and obtain dependable measurement accuracy for the life of those products.

Product Notes Time-Gated Spectrum Analysis (Agilent 8590-2) 5952-3685 Analog + Display 5091-2364E Maximizing Accuracy in Noise Figure Measurements (Agilent 85791-1) 5091-4801E

Application Notes Spectrum Analysis Basics (150) 5952-0292 Amplitude and Frequency Modulation (150)-1) 5954-9130

Get assistance with all your test and measurement needs at: www.agilent.com/find/assist

Product specifications and descriptions in this document subject to change without notice.

Copyright © 1996, 2000 Agilent Technologies Printed in U.S.A. 5/00 5963-6909E

Ask about the Agilent Technologies 8595E

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Agilent Technologies 8595E

Be the first to ask — other owners or our team will answer.

Original document

Agilent Technologies 8595E manual PDF

7 pages·179 KB·Water Dispenser·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…