Agilent Technologies 8360 B-Series Swept Signal Generator 8360 L-Series Swept CW Generator
Serial Number Prefixes:
This manual applies to any instrument with the following model number and serial number prefix combinations. You may have to modify this manual so that it applies directly to your instrument version. Refer to the “Instrument History” chapter.
Agilent Technologies 83640B/50B 4040A and Below
Agilent Technologies 83623L/30L 3844A and Below
Agilent Technologies 83640L/50L 4040A and Below

Printed in USA
October 2000
© Copyright 1997-2000 Agilent Technologies, Inc.

The material contained in this document is provided “as is”, and is subject to being changed, without notice, in future editions.
Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied with regard to this manual and to any of the Agilent products to which it pertains, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or any of the Agilent products to which it pertains. Should Agilent have a written contract with the User and should any of the contract terms conflict with these terms, the contract terms shall control.
We welcome any questions or comments you may have about our documentation. Please send us an E-mail at

Hewlett-Packard to Agilent Technologies Transition
This manual may contain references to HP or Hewlett-Packard. Please note that Hewlett- Packard's former test and measurement, semiconductor products and chemical analysis businesses are now part of Agilent Technologies. To reduce potential confusion, the only change to product numbers and names has been in the company name prefix: where a product name/number was HP XXXX the current name/number is now Agilent XXXX. For example, model number HP 8648 is now model number Agilent 8648.
The sales and service contact information in this manual may be out of date. The latest service and contact information for your location can be found on the Web at:
If you do not have access to the Internet, contact your field engineer. In any correspondence or telephone conversation, refer to your instrument by its model number and full serial number.


Agilent T ec hnologies certi es that this pro duct met its published sp eci cations at the time
of shipmen t from the factory . Agilent T ec hnologies further certi es that its calibration
measuremen ts are traceable to the United States National Institute of Standards and
T ec hnology , to the exten t allo w ed b y the Institute's calibration facility , and to the calibration
facilities of other In ternational Standards Organization mem b ers.
This Agilent T ec hnologies instrumen t pro duct is w arranted against defects in material and
w orkmanship for a p erio d of one y ear from date of shipmen t. During the w arrant y p erio d,
Agilen T ec hnologies Company will, at its option, either repair or replace pro ducts whic h prov e
to b e defectiv e.
F or w arran ty service or repair, this pro duct m ust b e returned to a service facilit y designated
b y Agilent T ec hnologies. Buy er shall prepa y shipping c harges to Agilent T ec hnologies and
Agilent T ec hnologies shall pa y shipping c harges to return the pro duct to Buy er. Ho w ev er,
Buy er shall pa y all shipping c harges, duties, and taxes for pro ducts returned to Agilent
T ec hnologies from another coun try .
Agilent T ec hnologies w arrants that its softw are and rm w are designated b y Agilent
T ec hnologies for use with an instrumen t will execute its programming instructions when
prop erly installed on that instrumen t. Agilent T ec hnologies do es not w arrant that the
op eration of the instrumen t, or softw are, or rm w are will b e unin terrupted or error-free.
Certification · W arranty

Limit a tion of W arranty
The foregoing w arrant y shall not apply to defects resulting from improp er or inadequate
maintenance b y Buy er, Buy er-supplied softw are or in terfacing, unauthorized mo di cation or
misuse, op eration outside of the en vironmental sp eci cations for the pro duct, or improp er
site preparation or maintenance.
NO OTHER W ARRANTY IS EXPRESSED OR IMPLIED. A GILENT TECHNOLOGIES
SPECIFICALL Y DISCLAIMS THE IMPLIED W ARRANTIES OF MER CHANT ABILITY
AND FITNESS FOR A P AR TICULAR PURPOSE.
THE REMEDIES PR O VIDED HEREIN ARE BUYER'S SOLE AND EX CLUSIVE
REMEDIES. A GILENT TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY
DIRECT, INDIRECT, SPECIAL, INCIDENT AL, OR CONSEQUENTIAL DAMA GES,
WHETHER BASED ON CONTRA CT, TOR T, OR ANY OTHER LEGAL THEOR Y.
Pr o duct maintenanc e agr e ements and other customer assistanc e agr e ements ar e available for
A gilent T e chnolo gies pr o ducts. F or any assistanc e, c ontact your ne ar est A gilent T e chnolo gies
Sales and Servic e Oc e.
c Cop yrigh t Agilent T ec hnologies 1996, 1997, 1999, 2000
All Righ ts Reserv ed. Repro duction, adaptation, or translation without prior written
p ermission is prohibited, except as allo w ed under the cop yright laws.
1400 F oun taingro v e P arkway , San ta Rosa CA, 95403-1799, USA
Exclusive Remedies · Assistance

The follo wing safety notes are used throughout this man ual. F amiliarize y ourself with eac h of
the notes and its meaning b efore op erating this instrumen t.
W arning denotes a hazard. It calls attention to a procedure which, if not
correctly performed or adhered to, could result in injury or loss of life. Do
not proceed bey ond a w arning note until the indicated conditions are fully
understood and met.
correctly p erformed or adhered to, w ould result in damage to or destruction
of the instrumen t. Do not pro ceed b ey ond a caution sign un til the indicated
conditions are fully understo o d and met.
Safety Notes

These servicing instructions are for use b y qualified personnel only . T o a v oid
electrical shock, do not perform any servicing unless y ou are qualified to do
so.
The opening of co v ers or remo v al of parts is lik ely to expose dangerous
v oltages. Disconnect the instrument from all v oltage sources while it is being
opened.
The detachable po w er cord is the instrument disconnecting device. It
disconnects the mains circuits from the mains supply before other parts of
the instrument. The front panel switch is only a standb y switch and is not a
LINE switch.
The po w er cord is connected to internal capacitors that may remain liv e for 5
seconds after disconnecting the plug from its po w er supply .
This is a Safety Class I product (pro vided with a protectiv e earthing ground
incorporated in the po w er cord). The mains plug shall only be inserted in a
sock et outlet pro vided with a protectiv e earth contact. Any interruption of the
protectiv e conductor, inside or outside the instrument, is lik ely to mak e the
instrument dangerous. Intentional interruption is prohibited.
For continued protection against fire hazard replace line fuse only with same
type and rating (F 5A/250V). The use of other fuses or material is prohibited.
General Safety Considerations

Always use the three-prong ac p ow er cord supplied with this instrumen t.
F ailure to ensure adequate earth grounding b y not using this cord may cause
instrumen t damage.
Before switching on this pro duct, make sure that the line v oltage selector
switch is set to the v oltage of the p ow er supply and the correct fuse is
installed. Assure the supply v oltage is in the sp eci ed range.
This pro duct is designed for use in Installation Category I I and P ollution
Degree 2 p er IEC 1010 and 664 resp ectively .
con v ection in to and out of the pro duct m ust not b e restricted. The ambien t
temp erature (outside the cabinet) m ust b e less than the maximum op erating
temp erature of the pro duct b y 4
C for ev ery 100 w atts dissipated in the
cabinet. If the total p ow er dissipated in the cabinet is greater than 800
w atts, then forced con v ection m ust b e used.
Before Applying P o w er
V erify that the pro duct is con gured to match the av ailable main p ow er source p er the input
p ow er con guration instructions provided in this man ual.
If this pro duct is to b e energized via an autotransformer make sure the common terminal is
connected to the neutral (grounded side of the mains supply).



This man ual provides service information for the Agilent T ec hnologies 8360 B-Series swept
signal generator and Agilent T ec hnologies 8360 L-Series swept CW generator.
This represen ts a k ey ph ysically lo cated on the instrumen t.
This indicates a \softkey ," a k ey whose lab el is determined b y the rm w are of
the instrumen t.
Screen Text This indicates text display ed on the instrumen t's screen.

Instruments Cov ered By This Manual
This man ual applies to instrumen ts ha ving a serial n um b er pre x listed on the title page
(b ehind the \Do cumen tation Map" tab). Some c hanges may ha v e to b e made to this man ual
so that it applies directly to eac h instrumen t; refer to Chapter 12, \Instrumen t History ," to
see what c hanges may apply to y our instrumen t.
A serial n um b er lab el ( Figure 0-1 ) is attached to the instrumen t's rear panel. A pre x (four
digits follo w ed b y a letter), and a sux ( v e digits unique to eac h instrumen t), comprise the
instrumen t serial n um b er.

Figure 0-1. Typical Serial Number Label
This man ual applies to the Agilent T ec hnologies 8360 B-Series swept signal generator and
Agilent T ec hnologies 8360 L-Series swept CW generator ha ving a mo del n um b er and serial
n um b er pre x listed on the title page (b ehind the \Do cumen tation Map" tab). Some c hanges
may ha v e to b e made to this man ual so that it applies directly to eac h instrumen t; refer to
Chapter 12, \Instrumen t History" in this man ual to see what c hanges may apply to y our
instrumen t.
This instrumen t has b een designed and tested in accordance with IEC Publication 1010,
Safety Requiremen ts for Electronic Measuring Apparatus, and has b een supplied in a safe
condition. The instruction do cumen tation con tains information and w arnings whic h m ust
b e follo w ed b y the user to ensure safe op eration and to maintain the instrumen t in a safe
condition.
Compliance

Sales and Service Offices
Hewlett-P ackard Sales and Service Offices
Instrument Supp ort Center
Headquarters F rance German y
Hewlett-Pac k ard S.A. Hewlett-Pac k ard F rance Hewlett-Pac k ard Gm bH
150, Route du Nan t-d'Avril 1 Av enue Du Canada Hewlett-Pac k ard Strasse
1217 Meyrin 2/Genev a Zone D'Activite De Courtab o euf 61352 Bad Hom burg v.d.H
Switzerland F-91947 Les Ulis Cedex German y
(41 22) 780.8111 F rance (49 6172) 16-0
(33 1) 69 82 60 60
Hewlett-Pac k ard Ltd.
Eskdale Road, Winnersh T riangle
W okingham, Berkshire R G41 5DZ
Hewlett-Pac k ard Company Hewlett-Pac k ard Australia Ltd. Hewlett-Pac k ard (Canada) Ltd.
3495 Deer Creek Road 31-41 Joseph Street 17500 South Service Road
P alo Alto, California, USA Blackburn, Victoria 3130 T rans- Canada High wa y
94304-1316 (61 3) 895-2895 Kirkland, Queb ec H9J 2X8 Dian District
Japan Singap ore T aiwan
Hewlett-Pac k ard Japan, Ltd. Hewlett-Pac k ard Singap ore (Pte.) Ltd. Hewlett-Pac k ard T aiwan
9-1 T ak akura-Cho, Hac hio ji 150 Beach Road 8th Flo or, H-P Building
T oky o 192, Japan #29-00 Gatew a y W est 337 F u Hsing North Road
(81 426) 60-2111 Singap ore 0718 T aip ei, T aiwan
(65) 291-9088 (886 2) 712-0404
China Hewlett-Pac k ard Company
38 Bei San Huan X1 Road
Shuang Y u Shu




In Case of Diculty 2-22
Description and Pro cedure 2-23
Related Adjustmen ts 2-24
In Case of Diculty 2-25
Description and Pro cedure 2-26
Lo w Band P ow er Flatness 2-26
High Band P ow er Flatness 2-27
Millimeter Band P ow er Flatness > 20 GHz and 40 GHz 2-27
Millimeter Band P ow er Flatness > 40 GHz 2-27
Related Adjustmen ts 2-27
In Case of Diculty 2-27
Description and Pro cedure 2-28
Related Adjustmen ts 2-29
In Case of Diculty 2-29
Description and Pro cedure 2-30
Related Adjustmen ts 2-31
In Case of Diculty 2-31
Description and Pro cedure 2-32
Harmonic Measuremen t < 20 GHz 2-33
Harmonic Measuremen t 20 to 26.5 GHz 2-35
Harmonic Measuremen t 26.5 to 40 GHz 2-37
Harmonic Measuremen t 38 to 50 GHz 2-37
Harmonic/Subharmonic V eri cation Pro cedure < 20 GHz 2-38
Harmonic/Subharmonic V eri cation Pro cedure > 20 GHz 2-39
Related Adjustmen ts 2-40
In Case of Diculty 2-40
Description and Pro cedure 2-41
Fixed Spurious Signals 2-41
Lo w Band O set Spurious Signals 2-43
Lo w Band Mixer Spurs 2-44
Related Adjustmen ts 2-44
In Case of Diculty 2-44
Description and Pro cedure 2-45
Related Adjustmen ts 2-47
In Case of Diculty 2-47
Description and Pro cedure 2-48
Related Adjustmen ts 2-51
In Case of Diculty 2-51
Description and Pro cedure 2-52
Related Adjustmen ts 2-54
In Case of Diculty 2-54
Contents-2 Agilent 8360 · Service

Description and Pro cedure 2-55
Rise and F all Times 2-55
Pulse Lev eling Accuracy 2-58
Related Adjustmen ts 2-59
In Case of Dicult y 2-59
Description and Pro cedure 2-60
Rise and F all Times < 20 GHz 2-60
Pulse Lev eling Accuracy < 20 GHz 2-64
Rise and F all Times > 20 GHz 2-67
Pro cedure > 20 to 26.5 GHz 2-68
Pro cedure > 26.5 to 40 GHz 2-68
Pro cedure > 40 GHz to 50 GHz 2-68
Pulse Lev eling Accuracy > 20 GHz 2-68
Pro cedure > 20 to 26.5 GHz 2-69
Pro cedure > 26.5 to 40 GHz 2-69
Pro cedure > 40 to 50 GHz 2-69
Related Adjustmen ts 2-70
In Case of Dicult y 2-70
Description and Pro cedure 2-71
System Calibration 2-73
Video Gain 2-73
Lo w Band Video F eedthrough 2-74
Video F eedthrough > 2 GHz 2-77
Related Adjustmen ts 2-77
In Case of Dicult y 2-77
Description and Pro cedure 2-78
Lo w Band < 2 GHz 2-78
F or Instrumen ts with Option 002 only 2-79
High Band > 2 GHz and < 20 GHz 2-80
F or Instrumen ts with Option 002 only 2-81
High Band > 20 GHz { Doubler Engaged 2-81
F or Instrumen ts with Option 002 only: 2-83
Related Adjustmen ts 2-83
In Case of Dicult y 2-83
Description and Pro cedure 2-84
Related Adjustmen ts 2-86
In Case of Dicult y 2-86
Description and Pro cedure 2-87
Normal AM Dynamic Range 2-88
Deep AM Dynamic Range 2-89
Related Adjustmen ts 2-91
In Case of Dicult y 2-91
Description and Pro cedure 2-92
External FM 2-92
In ternal FM (Option 002 only) 2-94
Contents-3 · Service

Related Adjustmen ts 2-94
In case of Diculty 2-94
Description and Pro cedure 2-95
Find Quadrature 2-98
DC FM Flatness (Unlo c k ed) 2-99
Related Adjustmen ts 2-99
In Case of Diculty 2-99
Description and Pro cedure 2-100
DC FM ( 100 Hz) (Using HP/Agilen t 8111A or HP/Agilent 8116A) . . 2-100
A C FM (at 100 kHz) (Using HP/Agilent 3325A) 2-102
Related Adjustmen ts 2-104
In Case of Diculty 2-104
Description and Pro cedure 2-105
Related Adjustmen ts 2-106
In Case of Diculty 2-106
Description and Pro cedure 2-107
Related Adjustmen ts 2-108
In Case of Diculty 2-108
Ho w T o Use This Chapter 3-1
Men u Maps 3-1
Equipmen t Required 3-1
List of Adjustmen ts 3-3
Description and Pro cedure 3-4
Related P erformance T ests 3-7
In Case of Diculty 3-7
Accuracy V ersus Adjustmen t In terv al 3-8
An Alternate Metho d 3-8
Description and Pro cedure 3-9
Related P erformance T ests 3-9
In Case of Diculty 3-9
Description and Pro cedure 3-10
200 MHz Lo op V CO T une Chec k 3-11
Sampler Match Adjustmen t 3-11
200 MHz Lo op Gain Adjustmen t 3-12
IF Gain Adjustmen t 3-12
Related P erformance T ests 3-12
In Case of Diculty 3-12
Description and Pro cedure 3-14
Related P erformance T ests 3-14
In Case of Diculty 3-14
Contents-4 Agilent 8360 · Service

Description and Pro cedure 3-15
Related P erformance T ests 3-16
In Case of Dicult y 3-16
Description and Pro cedure 3-17
Related P erformance T ests 3-17
In Case of Dicult y 3-17
Description 3-18
Adjustmen t Help 3-19
O set and Gain Adjustmen ts 3-19
Adjusting to the Cen ter of the P assband 3-19
SRD Bias Adjustmen t 3-19
Single Band Delay Comp ensation Adjustmen ts 3-19
Setting the F astest Sw eep Time 3-19
Sw eep Sp eed Related Adjustmen ts 3-19
YTM Bandcross Delay T erms 3-20
General T ec hniques 3-20
Pro cedure 3-21
YTM Gain Adjustmen t 3-25
Band 1 Adjustmen t 3-25
Band 2 Adjustmen t 3-26
Band 3 Adjustmen t 3-26
Band 4 Adjustmen t ( 83630B /L, 83640B /L, and 83650B /L only) . . . 3-27
Band 5 Adjustmen t ( 83640B /L and 83650B /L only) 3-28
Band 6 Adjustmen t ( 83640B /L and 83650B /L only) 3-29
Band 7 Adjustmen t ( 83650B /L only) 3-29
Multi-band Ampli er/Multiplier Delay Comp ensation 3-29
Band 4 Multi-band Ampli er/Multiplier Delay Comp ensation ( 83630B /L,
83640B /L, and 83650B /L only) 3-30
V eri cation 3-31
Related P erformance T ests 3-31
In Case Of Diculty 3-31
Description 3-32
Adjustmen t Help 3-33
O set and Gain Adjustmen ts 3-33
Squegg Clamp Calibration Constants 3-33
Adjusting for the Cen ter of the P assband 3-33
Single Band Delay Comp ensation Adjustmen ts 3-34
Setting the F astest Sw eep Time 3-34
Sw eep Sp eed Related Adjustmen ts 3-34
YTF Bandcross Delay T erms 3-34
Pro cedure 3-37
Ampli er/Filter O set and Gain Adjustmen t 3-41
Band 1 Adjustmen t 3-41
Band 2 Adjustmen t 3-43
Band 3 Adjustmen t 3-44
Band 4 Adjustmen t ( 83630B /L, 83640B /L, and 83650B /L only) . . . 3-46
Band 5 Adjustmen t ( 83640B /L and 83650B /L only) 3-47
Contents-5 · Service

Band 6 Adjustmen t ( 83640B /L and 83650B /L only) 3-48
Band 7 Adjustmen t ( 83650B /L only) 3-48
Auto T racking V eri cation 3-49
Single-Band Ampli er/Filter Delay 3-51
Band 1 Adjustmen t 3-52
Band 2 Adjustmen t 3-52
Band 3 Adjustmen t 3-53
Band 4 Adjustmen t ( 83630B /L, 83640B /L, and 83650B /L) 3-54
Band 5 Adjustmen t ( 83640B /L and 83650B /L only) 3-54
Band 6 Adjustmen t ( 83640B /L and 83650B /L only) 3-55
Band 7 Adjustmen t ( 83650B /L only) 3-55
Multi-band YTF Delay 3-56
Band 4 Multi-band YTF Delay ( 83630B /L, 83640B /L, and 83650B /L
only) 3-58
Op eration Chec k 3-59
Related P erformance T ests 3-59
In Case Of Diculty 3-59
Description and Pro cedure 3-60
Related P erformance T ests 3-61
In Case of Diculty 3-61
Description and Pro cedure 3-62
Related P erformance T ests 3-66
In Case of Diculty 3-66
Description and Pro cedure 3-67
Lo w Band ALC P ow er Accuracy Adjustmen t 3-68
High Band P ow er Accuracy Adjustmen t 3-69
Related P erformance T ests 3-70
In Case of Diculty 3-70
Description and Pro cedure 3-71
Lo w Band P ow er Flatness 3-71
High Band P ow er Flatness 3-72
Millimeter Band P ow er Flatness 3-72
Related P erformance T ests 3-72
In Case of Diculty 3-72
Description and Pro cedure 3-73
Related P erformance T ests 3-74
In Case of Diculty 3-74
Description and Pro cedure 3-75
AM O set 3-76
Log AM Accuracy 3-76
Linear AM Accuracy 3-77
Related P erformance T ests 3-77
In Case of Diculty 3-77
Description and Pro cedure 3-78
Contents-6 Agilent 8360 · Service

Description and Pro cedure 3-80
Related P erformance T ests 3-81
In Case of Dicult y 3-81
Description and Pro cedure 3-82
Related P erformance T ests 3-83
In Case of Dicult y 3-83
Description and Pro cedure 3-84
High Band Adjustmen t 3-84
Lo w Band Adjustmen t 3-85
Related P erformance T ests 3-85
In Case of Dicult y 3-85
Description and Pro cedure 3-86
A. A10 Jump er 3-87
B. A8 Jump er 3-88
Related P erformance T ests 3-88
In Case of Dicult y 3-88
Description and Pro cedure 3-89
A. A11 Jump er 3-90
B. A8 Jump er 3-91
Related P erformance T ests 3-91
In Case of Dicult y 3-91
Description and Pro cedure 3-92
Related P erformance T ests 3-93
In Case of Dicult y 3-93
Description and Pro cedure 3-94
T riggering on the Leading Edge 3-94
T riggering on the T railing Edge 3-94
Description and Pro cedure 3-95
Related P erformance T ests 3-95
In Case of Dicult y 3-96
Description and Pro cedure 3-97
Description and Pro cedure 3-98
Contents-7 · Service

In tro duction 4-1
De nition 4-1
Memory Areas 4-2
W orking Data 4-2
Protected Data 4-2
Default Data 4-2
Chec ksum V eri cation 4-2
Calibration Constant P assw ord 4-3
En tering a P assword 4-3
Disabling a P assword 4-3
Setting a P assword 4-4
Bypassing the P assword 4-5
Changing W orking Data Calibration Constants 4-6
Sa ving W orking Data Calibration Constants 4-6
Loading Protected Data Calibration Constants 4-7
Loading Default Data Calibration Constants 4-7
Calibration Constant Descriptions 4-8
In tro duction 5-1
Setting Up the System 5-2
Hardware Requiremen ts 5-2
Op erating System Requiremen ts 5-2
Installing the Softw are 5-3
Disk Files 5-3
Make W orking Copies 5-3
Installing the Program on to a Hard Disk 5-3
Op erating from a Floppy Disk 5-3
Running the Softw are 5-4
Con guration Limitations 5-4
Loading the Softw are 5-4
A UTOMA TED PERF ORMANCE TESTS 5-5
Description 5-5
Pro cedure 5-6
Description 5-7
Pro cedure 5-7
A UTOMA TED ADJUSTMENTS 5-10
Description 5-10
Pro cedure 5-10
Description 5-12
Description 5-13
Description 5-14
Pro cedure 5-14
Utilities 5-17
Contents-8 Agilent 8360 · Service

Description 5-17
Pro cedure 5-17
Description 5-20
Pro cedure 5-20
Description 5-20
Description 5-21
Loading Sensor Calibration F actors 5-21
T o Load Sensor Cal F actors: 5-21
Men u Ov erview 5-24
ALC Men u 6-3
F requency Men u 6-5
Marker Men u 6-7
Mo dulation Men u 6-9
P ow er Men u 6-11
Service Men u 6-13
Sw eep Men u 6-15
System Men u 6-17
User Cal Men u 6-19
In tro duction 7-1
T o ols Required 7-3
F ront P anel Disassembly and Reassem bly 7-4
T o ols Required 7-4
Description and Pro cedure 7-4
Rear P anel Disassembly and Reassem bly 7-21
T o ols Required 7-21
Description and Pro cedure 7-21
Contents-9 · Service

RF Dec k Disassem bly and Reassem bly 7-37
T o ols Required 7-37
Description and Pro cedure 7-37
Motherb oard Disassembly and Reassem bly 7-62
T o ols Required 7-62
Description and Pro cedure 7-62
In tro duction 9-1
Mo dule{Exc hange Program 9-1
Replacemen t Information 9-1
Ordering Information 9-2
Ho w to Order P arts . . . F ast! 9-2
In tro duction 10-1
Ho w to Clean the F an Filter 10-2
Ho w to Clean the Displa y Filter 10-4
Ho w to Replace the Line F use 10-5
Ho w to Replace the CPU Battery (A15BT1) 10-6
Remo v al/Replacemen t 10-6
Connector Care 10-7
Contents-10 Agilent 8360 · Service

In tro duction 11-1
Add A tten uator (Add Option 001) 11-2
Delete A tten uator (Delete Option 001) 11-2
Add Mo dulation/Generator (Add Option 002) 11-3
Rear P anel RF Output (Add Option 004) 11-4
F ront P anel RF Output (Delete Option 004) 11-4
1 Hz Capability (Add Option 008) 11-5
MA TE Compatibilit y (Add Option 700) 11-5
Rac k Moun t Slide Kit (Add Option 806) 11-6
Rac k Flange Kit Without Handles (Add Option 908) 11-6
Extra Manual Set (Add Option 910) 11-7
Rac k Flange Kit With Handles (Add Option 913) 11-7
Ho w to Use Instrumen t History 12-1
Change A 12-2
The A4 F ractional-N Assem bly has Changed 12-2
Replaceable P arts 12-2
Adjustmen ts 12-4
F ractional-N V CO Adjustmen t 12-4
Description and Pro cedure 12-4
Related P erformance T ests 12-4
In Case of Dicult y 12-5
F ractional-N Reference and API Spurs Adjustmen t 12-5
Description and Pro cedure 12-5
125 kHz Reference Spur Adjustmen t 12-6
API 1 Spur Adjustmen t 12-8
API 2 Spur Adjustmen t 12-8
Related P erformance T ests 12-8
In Case of Dicult y 12-8
Calibration Constants 12-9
P ost Repair 12-9
Index · Contents-11 · Service

0-1. T ypical Serial Num b er Lab el x
2-1. In ternal Timebase: Aging Rate T est Setup 2-8
2-2. Sw ept F requency Accuracy T est Setup 2-11
2-3. Video Signal on the Oscilloscop e 2-15
2-4. F requency Switching Time T est Setup 2-18
2-5. P ow er Accuracy T est Setup 2-23
2-6. P ow er Flatness T est Setup 2-26
2-7. Maximum Lev eled P ow er T est Setup 2-28
2-8. External Lev eling T est Setup 2-30
2-9. Spurious Signals (Harmonic) T est Setup < 20 GHz 2-32
2-10. Spurious Signals (Harmonics) T est Setup > 20 GHz 2-35
2-11. Spurious Signals (Non-Harmonic) T est Setup 2-41
2-12. Spurious signals (Line-Related) T est Setup 2-45
2-13. Single Sideband Phase Noise T est Setup 2-49
2-14. Pulse Mo dulation On/O Ratio T est Setup 2-52
2-15. Pulse P erformance T est Setup 2-55
2-16. Pulse En v elop e Display ed Ov er 5 Divisions 2-57
2-17. Widened Pulse En v elop e 2-58
2-18. Pulse P erformance T est Setup < 20 GHz 2-61
2-19. Signal Lev el b et w een RF P ow er O and On 2-64
2-20. Pulse P erformance T est Setup >20 GHz 2-67
2-21. Video F eedthrough T est Setup 2-71
2-22. Measuring Maximum V oltage from the Baseline Video F eedthrough 2-76
2-23. AM Accuracy T est Setup Lo w Band 2-78
2-24. AM Accuracy T est Setup < 20 GHz 2-80
2-25. AM Accuracy T est Setup > 20 GHz 2-82
2-26. AM Bandwidth T est Setup 2-84
2-27. AM Dynamic Range T est Setup 2-87
2-28. FM Accuracy T est Setup 2-92
2-29. FM Bandwidth T est Setup 2-97
2-30. Maximum FM Deviation T est Setup 2-100
2-31. Pulse Accuracy T est Setup 2-105
2-32. Mo dulation Meter T est Setup 2-107
3-1. 10 MHz Standard Adjustmen t Setup 3-5
3-2. 10 MHz Standard Adjustmen t Lo cation 3-6
3-3. Accuracy V ersus Adjustmen t In terv al 3-8
3-4. FL2 Lo cation 3-10
3-5. Sampler Match Adjustmen t Lo cations 3-11
3-6. IF Gain Adjustmen t Lo cation 3-12
3-7. YO Driv er +10 V Reference Adjustmen t Lo cation 3-14
3-8. YO Gain and Linearity Break P oints 3-15
3-9. YO Gain and Linearity Adjustmen t Lo cations 3-16
Figures · Contents-12 Agilent 8360 · Service

3-10. Switch and Adjustmen t Lo cations 3-17
3-11. Ampli er/Multiplier Adjustmen ts Setup 3-21
3-12. A12 Assem bly P oten tiometer Lo cations 3-27
3-16. Single-Band Delay and Risetime Comp ensation 3-35
3-17. Multi-Band Delay and Risetime Comp ensation 3-36
3-18. Ampli er/Filter Adjustmen ts Setup 3-37
3-19. XA10J2 Lo cation 3-38
3-20. A12 Assem bly P oten tiometer Lo cations 3-45
3-21. Lo w P ow er SRD Bias Adjustmen t Setup 3-60
3-22. A9R105 Lo cation 3-61
3-23. Mo dulator O set and Gain Adjustmen t Setup 3-62
3-24. ALC P ow er Lev el Accuracy Adjustmen t Setup 3-67
3-25. A10R125 Lo cation 3-69
3-26. P ow er Flatness Adjustmen t Setup 3-71
3-27. A8R111 and A8R105 Adjustmen t Lo cations 3-73
3-28. AM Accuracy Adjustmen t Setup 3-75
3-29. A10R35 and A10R39 Lo cations 3-77
3-30. AM O set Adjustmen t Setup 3-79
3-31. AM Delay Adjustmen t Setup 3-80
3-32. AM Delay Adjustmen t Lo cation 3-81
3-33. FM Gain Adjustmen t Setup 3-82
3-34. FM Gain Adjustmen t Lo cation 3-83
3-35. Square W av e Symmetry Adjustmen t Setup 3-84
3-36. A10 Jump er Lo cations 3-87
3-37. A8 Jump er Lo cations 3-88
3-38. FM Input Imp edance Jump er Lo cations 3-90
3-39. A8 Jump er Lo cations 3-91
3-40. Mo dulation Generator Flatness Adjustmen t Setup 3-92
3-41. External T rigger Edge Selection Lo cation 3-94
3-42. A12S1 Lo cation 3-95
5-1. Step A tten uator Flatness T est Setup 5-6
5-2. Step A tten uator Flatness Adjustmen t Setup 5-11
6-1. ALC Men u 6-3
6-2. F requency Men u 6-5
6-3. Marker Men u 6-7
6-4. Mo dulation Men u 6-9
6-5. P ow er Men u 6-11
6-6. Service Men u 6-13
6-7. Sw eep Men u 6-15
6-8. System Men u 6-17
6-9. User Cal Men u 6-19
7-1. Cable Routing Lo cations 7-3
7-2. F ront P anel Disassem bly Flow c hart 7-5
7-3. F ront Edge T rim Strip 7-6
7-4. F ront P anel Screw Lo cations 7-7
7-5. F ront P anel Coaxial and Ribb on Cable Lo cations 7-8
7-6. F ront P anel Pro cessor Cables 7-9
7-7. Displa y Remo v al 7-10
7-8. Source Mo dule In terface Remo v al 7-11
7-9. RPG Remo v al 7-12
7-10. Remo ving RF Output Connector (F ront P anel) 7-13
Contents-13 · Service

7-11. F ront P anel BNC Lo cations 7-14
7-12. Keyb oard Assem bly Remo v al 7-15
7-13. Keypad and An ti-ro c k Sheets 7-16
7-14. Display Filter 7-17
7-15. Keyb oard Reassem bly 7-18
7-16. Keyb oard Assem bly Screw Lo cations 7-19
7-17. Rear P anel Disassembly Flo w c hart 7-22
7-18. B1 Rear P anel F an Assem bly 7-24
7-19. Rear P anel BNC Connector Lo cations 7-25
7-20. Remo ving RF Output Connector 7-26
7-21. Rear P anel Remo v al 7-27
7-22. Rear P anel Cable Lo cations 7-28
7-23. Reference Standard Moun tings 7-29
7-24. T ransformer Remo v al 7-30
7-25. A19 Rear P anel In terface Remo v al 7-31
7-26. V oltage Selection Switch Remo v al 7-32
7-27. F use, Line Filter, and Ground Wires 7-33
7-28. Rear P anel F use Housing 7-34
7-29. Line Filter Remo v al 7-35
7-30. Option 004 BNC Connectors Remo v al 7-36
7-31. RF Dec k Disassembly Flow c hart 7-38
7-32. Remo ving RF Output Connector 7-40
7-33. W51 RF Cable Lo cation 7-41
7-34. RF Dec k in the Service P osition 7-42
7-35. A20 RF In terface 7-43
7-36. RF Dec k Cable Lo cations { 83640B /L and 83650B /L 7-45
7-37. RF Dec k Cable Lo cations { 83640B /L Option 001 and 83650B /L Option 001 7-46
7-38. RF Dec k Cable Lo cations { 83640B /50B Option 006 7-47
7-39. RF Dec k Cable Lo cations { 83640B /50B Options 001 and 006 7-48
7-40. RF Dec k Cable Lo cations { 83620B /22B and 83630B /L 7-49
7-41. RF Dec k Cable Lo cations { 83620B /22B and 83630B /L Option 001 7-50
7-42. RF Dec k Cable Lo cations { 83620B /22B/30B Option 006 7-51
7-43. RF Dec k Cable Lo cations { 83620B /22B/30B Options 001 and 006 7-52
7-44. RF Dec k Cable Lo cations { 83623B 7-53
7-45. RF Dec k Cable Lo cations { 83623B Option 001 7-54
7-46. RF Dec k Cable Lo cations { 83624B 7-57
7-47. RF Dec k Cable Lo cations { 83624B Option 001 7-58
7-48. RF Dec k Cable Lo cations { 83623B /24B Option 006 7-59
7-49. RF Dec k Cable Lo cations { 83623B /24B Options 001 and 006 7-60
7-50. RF Dec k Screw Lo cations 7-61
7-51. Motherb oard Disassembly Flow c hart 7-63
7-52. F ront Edge T rim Strip 7-64
7-53. F ront P anel Screw Lo cations 7-65
7-54. F ront P anel Coaxial and Ribb on Cable Lo cations 7-66
7-55. Remo ving RF Output Connector (F ront P anel) 7-67
7-56. Remo ving RF Output Connector (Option 004) 7-69
7-57. Rear P anel Remo v al 7-70
7-58. Rear P anel Cable Lo cations 7-71
7-59. Ma jor Assem blies 7-72
7-60. W51 Lo cation 7-73
7-61. Motherb oard Cables 7-74
Contents-14 Agilent 8360 · Service

7-62. Motherb oard Screws 7-75
9-1. Mo dule Exc hange Program 9-4
9-2. Ma jor Assem blies (1 of 5) 9-5
9-2. Ma jor Assem blies (2 of 5) 9-6
9-2. Ma jor Assem blies (3 of 5) 9-7
9-2. Ma jor Assem blies (4 of 5) 9-8
9-2. Ma jor Assem blies (5 of 5) 9-9
9-3. Cables (1 of 19) 9-15
9-3. Cables (2 of 19) 9-16
9-3. Cables (3 of 19) 9-17
9-3. Cables (4 of 19) 9-18
9-3. Cables (5 of 19) 9-19
9-3. Cables (6 of 19) 9-20
9-3. Cables (7 of 19) 9-21
9-3. Cables (8 of 19) 9-22
9-3. Cables (9 of 19) 9-23
9-3. Cables (10 of 19) 9-24
9-3. Cables (11 of 19) 9-25
9-3. Cables (12 of 19) 9-26
9-3. Cables (13 of 19) 9-27
9-3. Cables (14 of 19) 9-28
9-3. Cables (15 of 19) 9-29
9-3. Cables (16 of 19) 9-30
9-3. Cables (17 of 19) 9-31
9-3. Cables (18 of 19) 9-32
9-3. Cables (19 of 19) 9-33
9-4. F ront P anel Hardware 9-40
9-5. F ront P anel, Casting and Keypad 9-42
9-6. F ront P anel, A ttaching Hardware (1 of 3) 9-43
9-6. F ront P anel, A ttaching Hardware (2 of 3) 9-44
9-6. F ront P anel, A ttaching Hardware (3 of 3) 9-45
9-7. Rear P anel Hardware 9-46
9-8. Rear P anel F an Assem bly 9-49
9-9. Bottom View, A ttaching Hardware 9-50
9-10. T op View, A ttaching Hardware 9-51
9-11. Left Side View, A ttaching Hardware 9-53
9-12. Righ t Side View, A ttaching Hardware 9-54
9-13. RF Dec k Righ t Side View, A ttaching Hardware 9-55
9-14. RF Connector and A ttaching Hardware 9-56
9-15. Chassis P arts 9-57
10-1. Remo ving the F an Filter 10-3
10-2. Replacing the Line F use 10-5
10-3. A15BT1 Battery Lo cation 10-6
12-3. Ma jor Assem bly and Cable Lo cations 12-3
12-4. A4L1 Lo cation 12-4
12-5. F ractional-N Reference and API Spur Adjustmen t Setup 12-6
12-6. F ractional-N Reference and API Spurs Adjustmen t Lo cations 12-7
Contents-15 · Service

2-1. Op eration V eri cation F orm 2-4
2-2. Sw ept F requency Accuracy Instrumen t Settings 2-14
2-3. Additional Instrumen t Settings 2-15
2-4. Additional Instrumen t Settings 2-16
2-5. Additional Instrumen t Settings 2-16
2-6. Additional Instrumen t Settings 2-17
2-7. CW F requency Switching Time Settings 2-19
2-8. P ow er Accuracy F requencies 2-24
2-9. Start and Stop F requencies 2-33
2-10. Corresp onding YO F requency Ranges and RF Output F requencies 2-33
2-11. Corresp onding YO F requency Ranges and RF Output F requencies 2-36
2-12. Determining the Actual Spur F requencies 2-42
2-13. Lo w Band Spurious Signals 2-43
2-14. Sp ectrum Analyzer F requencies and Line Spur F requencies and Harmonics . 2-46
2-15. Phase Noise Measuremen t System P arameters 2-50
2-16. F requency Setting for Phase Noise Measuremen ts 2-51
2-17. Pulse On/O Cen ter F requencies 2-53
2-18. Syn thesizer F requencies 2-56
2-19. Syn thesizer and LO F requencies 2-62
2-20. Risetime and F alltime V alues 2-68
2-21. Lo w Band Video F eedthrough F requencies 2-74
2-22. P ow er Lev el Con v ersions to V olts P eak 2-76
2-23. Millimeter Band Video F eedthrough F requencies 2-77
2-24. Syn thesizer F requencies 2-85
2-25. P ow er Meter Readings 2-89
2-26. RF Signal Lev el 2-90
2-27. Syn thesizer F requencies 2-93
2-28. Syn thesizer F requencies 2-98
2-29. DC FM T est P ass-F ail Results 2-101
2-30. FM Deviation F requencies and Settings 2-103
2-33. T est Record for Agilent 83620B and 83622B (1 of 8) 2-109
2-33. T est Record for 83620B and 83622B (2 of 8) 2-110
2-33. T est Record for 83620B and 83622B (3 of 8) 2-111
2-33. T est Record for 83620B and 83622B (4 of 8) 2-112
2-33. T est Record for 83620B and 83622B (5 of 8) 2-113
2-33. T est Record for 83620B and 83622B (6 of 8) 2-114
2-33. T est Record for 83620B and 83622B (7 of 8) 2-115
2-33. T est Record for 83620B and 83622B (8 of 8) 2-116
2-34. T est Record for Agilent 83623B and 83624B (1 of 8) 2-117
2-34. T est Record for 83623B and 83624B (2 of 8) 2-118
2-34. T est Record for 83623B and 83624B (3 of 8) 2-119
2-34. T est Record for 83623B and 83624B (4 of 8) 2-120
2-34. T est Record for 83623B and 83624B (5 of 8) 2-121
T ables · Contents-16 Agilent 8360 · Service

2-34. T est Record for 83623B and 83624B (6 of 8) 2-122
2-34. T est Record for 83623B and 83624B (7 of 8) 2-123
2-34. T est Record for 83623B and 83624B (8 of 8) 2-124
2-35. T est Record for Agilent 83630B (1 of 8) 2-125
2-35. T est Record for 83630B (2 of 8) 2-126
2-35. T est Record for 83630B (3 of 8) 2-127
2-35. T est Record for 83630B (4 of 8) 2-128
2-35. T est Record for 83630B (5 of 8) 2-129
2-35. T est Record for 83630B (6 of 8) 2-130
2-35. T est Record for 83630B (7 of 8) 2-131
2-35. T est Record for 83630B (8 of 8) 2-132
2-36. T est Record for Agilent 83640B (1 of 8) 2-133
2-36. T est Record for 83640B (2 of 8) 2-134
2-36. T est Record for 83640B (3 of 8) 2-135
2-36. T est Record for the 83640B (4 of 8) 2-136
2-36. T est Record for the 83640B (5 of 8) 2-137
2-36. T est Record for the 83640B (6 of 8) 2-138
2-36. T est Record for the 83640B (7 of 8) 2-139
2-36. T est Record for the 83640B (8 of 8) 2-140
2-37. T est Record for Agilent 83650B (1 of 8) 2-141
2-37. T est Record for 83650B (2 of 8) 2-142
2-37. T est Record for 83650B (3 of 8) 2-143
2-37. T est Record for 83650B (4 of 8) 2-144
2-37. T est Record for 83650B (5 of 8) 2-145
2-37. T est Record for 83650B (6 of 8) 2-146
2-37. T est Record for 83650B (7 of 8) 2-147
2-37. T est Record for 83650B (8 of 8) 2-148
2-34. T est Record for Agilent 83623L (1 of 5) 2-149
2-34. T est Record for 83623L (2 of 5) 2-150
2-34. T est Record for 83623L (3 of 5) 2-151
2-34. T est Record for 83623L (4 of 5) 2-152
2-34. T est Record for 83623L (5 of 5) 2-153
2-35. T est Record for Agilent 83630L (1 of 5) 2-154
2-35. T est Record for 83630L (2 of 5) 2-155
2-35. T est Record for 83630L (3 of 5) 2-156
2-35. T est Record for 83630L (4 of 5) 2-157
2-35. T est Record for 83630L (5 of 5) 2-158
2-36. T est Record for Agilent 83640L (1 of 5) 2-159
2-36. T est Record for 83640L (2 of 5) 2-160
2-36. T est Record for 83640L (3 of 5) 2-161
2-36. T est Record for the 83640L (4 of 5) 2-162
2-36. T est Record for the 83640L (5 of 5) 2-163
2-37. T est Record for Agilent 83650L (1 of 6) 2-164
2-37. T est Record for 83650L (2 of 6) 2-165
2-37. T est Record for 83650L (3 of 6) 2-166
2-37. T est Record for 83650L (4 of 6) 2-167
2-37. T est Record for 83650L (5 of 6) 2-168
2-37. T est Record for 83650L (6 of 6) 2-169
3-1. Calibration Constant Descriptions 3-22
3-2. Start and Stop F requencies (GHz) 3-31
3-4. Ampli er/Filter Calibration Constants and Default V alues 3-39
Contents-17 · Service

3-5. F requency Ranges 3-50
3-6. ALC O set Calibration Constant Default V alues 3-63
3-7. Cen ter F requencies and Calibration Constants 3-65
3-8. ALC Calibration Constants and Default V alues 3-68
3-9. P ow er Lev el and Calibration Constant Adjustmen t 3-69
3-10. P ow er Lev el and Calibration Constant Adjustmen t 3-70
4-9. Calibration Constant Descriptions 4-9
4-10. Hardware Con guration Cal Constants #463 and #464 4-32
4-11. Assem blies and Corresp onding Cal Constant ID Num b ers and Revisions . . 4-33
5-1. Required Binaries 5-2
5-2. T est Equipmen t Required 5-5
5-3. T est Equipmen t Required 5-10
5-4. Syn thesizer Hardk ey Em ulation 5-18
8-1. Adjustmen ts and P erformance T ests Required After Repair or Replacemen t of
an Assem bly 8-1
9-1. Reference Designations and Abbreviations 9-3
9-2. Ma jor Assem blies 9-10
9-3. Cables 9-34
9-4. F ront P anel Hardware 9-41
9-5. F ront P anel, Casting and Keypad 9-42
9-6. F ront P anel, A ttaching Hardware 9-43
9-7. Rear P anel (1 of 2) 9-47
9-7. Rear P anel (2 of 2) 9-48
9-8. Rear P anel F an Assem bly 9-49
9-9. Bottom View, A ttac hing Hardware 9-50
9-10. T op View, A ttaching Hardware 9-52
9-11. Left Side View, A ttaching Hardware 9-53
9-12. Righ t Side View, A ttac hing Hardware 9-54
9-13. RF Dec k Righ t Side View, A ttaching Hardware 9-55
9-14. RF Connector and A ttaching Hardware 9-56
9-15. Chassis P arts 9-58
9-16. Miscellaneous Replaceable Accessories 9-59
12-1 12-2
12-2. Summary of F ractional-N Reference & API Spurs Adjustmen t 12-5
12-3. Assem blies and Corresp onding Cal Constant ID Num b ers and Revisions . . 12-9
12-4. Adjustmen ts and P erformance T ests Required After Repair or Replacemen t of
an Assem bly 12-9
Contents-18 Agilent 8360 · Service

Eac h R e quir e d Equipment list is sho wn b elow. The page n um b er of eac h list is also included
for easy access.
Agilent 83620B /22B Required Equipmen t : 1-2
Agilent 83623B /24B Required Equipmen t : 1-9
Agilent 83630B Required Equipmen t : 1-16
Agilent 83640B Required Equipmen t : 1-23
Agilent 83650B Required Equipmen t : 1-30
Agilent 83623L Required Equipmen t : 1-38
Agilent 83630L Required Equipmen t : 1-43
Agilent 83640L Required Equipmen t : 1-48
Agilent 83650L Required Equipmen t : 1-53
Agilent 8360 Required Equipment
Equipment Required · Agilent 83620B/22B

Agilent 83620B/22B Required Equipment
T able 1-1 Required Equipment for 83620B/22B
Instrument Critical Sp eci cations Recommended
80 dB display (1.0 dB maxim um)
to v ertical p osition of trace on display
Pulse Mo dulation On/O
Capable of phase lo cking to external
AM Dynamic Range (P)
Maxim um FM Deviation (P)
and API Spurs (A)
Square W a v e Symmetry (A)
A vs B Sweep Mo de
Pulse P erformance-Alt (P)
Controller 4 Mb yte RAM
Step A ttenuator Flatness (AP)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.

T able 1-1. Required Equipment for 83620B/22B (continued)
Instrument Critical Sp eci cations Recommended
P/N 08360-10001 Step A ttenuator Flatness (AP)
(Shipp ed with instrument)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
AM Dynamic Range (P)
Mo dulation Meter (P)
Lo w P o wer SRD Bias (A)
F requency Switc hing Time (P)
Pulse P erformance-Alt (P)
Pulse Mo dulation Video
In ternal Pulse Accuracy (P)
10 MHz Standard (A)
Mo dulator O set and Gain (A)
Pulse P erformance (P)
scale 2 mV
T rigger Sensitivit y DC to 100 MHz:
T rigger & Timebase Jitter:
2.5 ps + 5E
Time In terv al Accuracy:
10 ps 0.1% of reading
Channel Input Re ection:
5% for 30 ps risetime
1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.
Agilent 8360 Required Equipment

T able 1-1. Required Equipment for 83620B/22B (continued)
Instrument Critical Sp eci cations Recommended
F requency Switching Time (P)
Pulse Mo dulation Video
Mo dulator O set and Gain (A)
Pulse P erformance (P)
Pulse Mo dulation Video
HP/Agilent 3325A/B AM Accuracy (P)
AM Dynamic Range (P)
Maxim um FM Deviation (P)
Mo dulation Meter (P)
Maxim um FM Deviation (P)
P o wer Meter P o wer Range: 1 W to 100 mW
P o wer Accuracy (P)
AM Dynamic Range (P)
ALC P o wer Level
P o wer Meter P o wer Range: 1 W to 100 mW
HP/Agilent 437B P o wer Flatness (P)
P o wer Flatness (A)
1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.

T able 1-1. Required Equipment for 83620B/22B (continued)
Instrument Critical Sp eci cations Recommended
P o wer Meter P o wer Range: 1 W to 100 mW
HP/Agilen t 438A P o wer Flatness and
P o wer Flatness (AA)
Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
P o wer Sensor F requency Range: 10 MHz to 2.3 GHz
HP/Agilen t 8482A P o wer Flatness (P)
P o wer Flatness (A)
Step A ttenuator Flatness (AP)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
P o wer Sensor F requency Range: 50 MHz to 20 GHz
HP/Agilen t 8485D Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
P o wer Sensor F requency Range: 50 MHz to 20 GHz
HP/Agilen t 8485A P o wer Accuracy (P)
P o wer Flatness (P)
AM Dynamic Range (P)
ALC P o wer Level
P o wer Flatness (A)
P o wer Flatness and
P o wer Flatness (AA)
Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
A ttenuator A ttenuation: 30 dB 0.05 dB at 50 MHz HP/Agilen t 11708A Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
F requency Range (tuned):
2.5 MHz to 1.3 GHz
1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.
Agilent 8360 Required Equipment

T able 1-1. Required Equipment for 83620B/22B (continued)
Instrument Critical Sp eci cations Recommended
F requency Range (carrier):
HP/Agilent 3048A Single Sideband Phase
0.01 to 18 GHz
Sensitivit y:
< 70 dBc at 100 Hz o set
< 78 dBc at 1 kHz o set
< 86 dBc at 10 kHz o set
< 107 dBc at 100 kHz o set
O set F requency Range:
100 Hz to 2 MHz
Amplitude Accuracy:
2 dB to 1 MHz o set
10 MHz Standard (A)
HP/Agilent 8348A FM Bandwidth (P)
HP/Agilent 8447F Pulse P erformance-Alt (P)
Pulse Mo dulation Video
P o wer Ampli er Gain: 22 dB
1 meter of coax or
P/N 0955-0307 Pulse P erformance-Alt (P)
P o wer Splitter F requency Range: 10 MHz to 20 GHz HP/Agilent 11667B FM Bandwidth (P)
HP/Agilent 33330D External Leveling (P)
Pulse P erformance-Alt (P)
Maxim um Input P o wer: 300 mW
Pulse Mo dulation Video
1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est
P { Manual P erformance T est

T able 1-1. Required Equipment for 83620B/22B (continued)
Instrument Critical Sp eci cations Recommended
P o wer Accuracy (P)
Maxim um Input P o wer: 300 mW
Pulse P erformance (P)
Pulse P erformance-Alt (P)
FM Bandwidth (P) (2)
Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AA)
Maxim um Leveled P o wer (P)
Maxim um Input P o wer: 300 mW
Pulse P erformance (P)
3.7 GHz Lo w
P/N 9135-0191 Pulse Mo dulation Video
K & L Microwa v e
Pulse Mo dulation Video
500 MHz Lo w
Pulse Mo dulation Video
T o ol Kit No Substitute P/N 08360{60060
In v ertron California
Capacitor 1000 pf P/N 0160-4574 External Leveling (P)
1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est
P { Manual P erformance T est
Agilent 8360 Required Equipment

The following list of adapters and cables is provided for con v enience. They may b e used in
equipmen t setups for p erformance tests or adjustmen ts.
SMA (m) to SMA (m) adapter 1250-1159
SMA (f ) to SMA (f ) adapter 1250-1158
SMB (m) to SMB (m) adapter 1250-0669
SMB (f ) to SMB (f ) adapter 1250-0672
SMB tee (f ) (m) (m) 1250-1391
3.5 mm (f ) to 3.5 mm (f ) adapter 5061-5311
3.5 mm (f ) to N-t yp e (m) adapter 1250-1745
2.4 mm (f ) to 2.92 mm (f ) adapter 1250-2187
2.4 mm (f ) to 2.4 mm (f ) adapter 1250-2188
2.4 mm (m) to 3.5 mm (f ) adapter 11901C
BNC (f ) to BNC (f ) adapter 1250-0080
BNC (m) to BNC (m) adapter 1250-0216
BNC (f ) to SMA (m) adapter 1250-1200
BNC (f ) to SMB (m) adapter 1250-1237
BNC tee (m) (f ) (f ) 1250-0781
SMB (f ) to BNC (m) exible cable 85680-60093
SMA semi-rigid cable 2 feet 08340-20124
GPIB cable 2 meters HP/Agilent 10833B
BNC cable 8120-2582 · 1-8 Required Equipment · Agilent 83620B/22B

Agilent 83623B/24B Required Equipment
T able 1-2. Required Equipment for 83623B/24B
Instrument Critical Sp eci cations Recommended
80 dB display (1.0 dB maxim um)
to v ertical p osition of trace on display
Pulse Mo dulation On/O
Capable of phase lo cking to external
AM Dynamic Range (P)
Maxim um FM Deviation (P)
and API Spurs (A)
Square W a v e Symmetry (A)
A vs B Sweep Mo de
Pulse P erformance-Alt (P)
Controller 4 Mb yte RAM
Step A ttenuator Flatness (AP)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.
Agilent 8360 Required Equipment

T able 1-2. Required Equipment for 83623B/24B (continued)
Instrument Critical Sp eci cations Recommended
P/N 08360-10001 Step A ttenuator Flatness (AP)
(Shipp ed with instrument)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
AM Dynamic Range (P)
Mo dulation Meter (P)
Lo w P o wer SRD Bias (A)
F requency Switching Time (P)
Pulse P erformance-Alt (P)
Pulse Mo dulation Video
In ternal Pulse Accuracy (P)
10 MHz Standard (A)
Mo dulator O set and Gain (A)
Pulse P erformance (P)
scale 2 mV
T rigger Sensitivit y DC to 100 MHz:
T rigger & Timebase Jitter:
2.5 ps + 5E
Time In terv al Accuracy:
10 ps 0.1% of reading
Channel Input Re ection:
5% for 30 ps risetime
1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.

T able 1-2. Required Equipment for 83623B/24B (continued)
Instrument Critical Sp eci cations Recommended
F requency Switc hing Time (P)
Pulse Mo dulation Video
Mo dulator O set and Gain (A)
Pulse P erformance (P)
Pulse Mo dulation Video
HP/Agilen t 3325A/B AM Accuracy (P)
AM Dynamic Range (P)
Maxim um FM Deviation (P)
Mo dulation Meter (P)
Maxim um FM Deviation (P)
P o wer Meter P o wer Range: 1 W to 100 mW
P o wer Accuracy (P)
AM Dynamic Range (P)
ALC P o wer Level
P o wer Meter P o wer Range: 1 W to 100 mW
HP/Agilen t 437B P o wer Flatness (P)
P o wer Flatness (A)
P o wer Meter P o wer Range: 1 W to 100 mW
HP/Agilen t 438A P o wer Flatness and
P o wer Flatness (AA)
Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.
Agilent 8360 Required Equipment

T able 1-2. Required Equipment for 83623B/24B (continued)
Instrument Critical Sp eci cations Recommended
P o wer Sensor F requency Range: 10 MHz to 2.3 GHz
HP/Agilent 8482A P o wer Flatness (P)
P o wer Flatness (A)
Step A ttenuator Flatness (AP)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
P o wer Sensor F requency Range: 50 MHz to 20 GHz
HP/Agilent 8485D Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
P o wer Sensor F requency Range: 50 MHz to 20 GHz
HP/Agilent 8485A P o wer Accuracy (P)
P o wer Flatness (P)
AM Dynamic Range (P)
ALC P o wer Level
P o wer Flatness (A)
P o wer Flatness and
P o wer Flatness (AA)
Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
A ttenuator A ttenuation: 30 dB 0.05 dB at 50 MHz HP/Agilent 11708A Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
F requency Range (tuned):
2.5 MHz to 1.3 GHz
F requency Range (carrier):
HP 3048A Single Sideband Phase
0.01 to 18 GHz
Sensitivit y:
< 70 dBc at 100 Hz o set
< 78 dBc at 1 kHz o set
< 86 dBc at 10 kHz o set
< 107 dBc at 100 kHz o set
O set F requency Range:
100 Hz to 2 MHz
Amplitude Accuracy:
2 dB to 1 MHz o set
1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.

T able 1-2. Required Equipment for 83623B/24B (continued)
Instrument Critical Sp eci cations Recommended
10 MHz Standard (A)
HP/Agilen t 8348A FM Bandwidth (P)
HP/Agilen t 8447F Pulse P erformance-Alt (P)
Pulse Mo dulation Video
P o wer Ampli er Gain: 22 dB
1 meter of coax or
P/N 0955-0307 Pulse P erformance-Alt (P)
P o wer Splitter F requency Range: 10 MHz to 20 GHz HP/Agilen t 11667B FM Bandwidth (P)
HP/Agilen t 33330D External Leveling (P)
Pulse P erformance-Alt (P)
Maxim um Input P o wer: 300 mW
Pulse Mo dulation Video
P o wer Accuracy (P)
Maxim um Input P o wer: 300 mW
Pulse P erformance (P)
Pulse P erformance-Alt (P)
FM Bandwidth (P) (2)
Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AA)
Maxim um Leveled P o wer (P)
Maxim um Input P o wer: 300 mW
Pulse P erformance (P)
1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est
P { Manual P erformance T est
Agilent 8360 Required Equipment

T able 1-2. Required Equipment for 83623B/24B (continued)
Instrument Critical Sp eci cations Recommended
3.7 GHz Lo w
P/N 9135-0191 Pulse Mo dulation Video
K & L Microwa v e
Pulse Mo dulation Video
500 MHz Lo w
Pulse Mo dulation Video
T o ol Kit No Substitute P/N 08360{60060
In v ertron California
Capacitor 1000 pf P/N 0160-4574 External Leveling (P)
1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est
P { Manual P erformance T est

The follo wing list of adapters and cables is provided for con v enience. They may b e used in
equipmen t setups for p erformance tests or adjustmen ts.
SMA (m) to SMA (m) adapter 1250-1159
SMA (f ) to SMA (f ) adapter 1250-1158
SMB (m) to SMB (m) adapter 1250-0669
SMB (f ) to SMB (f ) adapter 1250-0672
SMB tee (f ) (m) (m) 1250-1391
3.5 mm (f ) to 3.5 mm (f ) adapter 5061-5311
3.5 mm (f ) to N-t yp e (m) adapter 1250-1745
2.4 mm (f ) to 2.92 mm (f ) adapter 1250-2187
2.4 mm (f ) to 2.4 mm (f ) adapter 1250-2188
2.4 mm (m) to 3.5 mm (f ) adapter 11901C
BNC (f ) to BNC (f ) adapter 1250-0080
BNC (m) to BNC (m) adapter 1250-0216
BNC (f ) to SMA (m) adapter 1250-1200
BNC (f ) to SMB (m) adapter 1250-1237
BNC tee (m) (f ) (f ) 1250-0781
SMB (f ) to BNC (m) exible cable 85680-60093
SMA semi-rigid cable 2 feet 08340-20124
GPB cable 2 meters HP/Agilent 10833B
Agilent 8360 Required Equipment
BNC cable 8120-2582 · Agilent 83623B/24B

Agilent 83630B Required Equipment
T able 1-3. Required Equipment for 83630B
Instrument Critical Sp eci cations Recommended
(26.5 GHz with external mixers)
80 dB display (1.0 dB maxim um)
Pulse Mo dulation On/O
to v ertical p osition of trace on display
Capable of phase lo cking to external
Pulse P erformance-Alt (P)
AM Dynamic Range (P)
Maxim um FM Deviation (P)
and API Spurs (A)
Square W a v e Symmetry (A)
A vs B Sweep Mo de
Pulse P erformance-Alt (P)
Controller 4 Mb yte RAM
Step A ttenuator Flatness (AP)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.

T able 1-3. Required Equipment for 83630B (continued)
Instrument Critical Sp eci cations Recommended
P/N 08360-10001 Step A ttenuator Flatness (AP)
(Shipp ed with instrument)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
AM Dynamic Range (P)
Mo dulation Meter (P)
Lo w P o wer SRD Bias (A)
F requency Switc hing Time (P)
Pulse P erformance-Alt (P)
Pulse Mo dulation Video
In ternal Pulse Accuracy (P)
10 MHz Standard (A)
Mo dulator O set and Gain (A)
Pulse P erformance (P)
scale 2 mV
T rigger Sensitivit y DC to 100 MHz:
T rigger & Timebase Jitter:
2.5 ps + 5E
Time In terv al Accuracy:
10 ps 0.1% of reading
Channel Input Re ection:
5% for 30 ps risetime
1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.
Agilent 8360 Required Equipment

T able 1-3. Required Equipment for 83630B (continued)
Instrument Critical Sp eci cations Recommended
F requency Switching Time (P)
Pulse Mo dulation Video
Mo dulator O set and Gain (A)
Pulse P erformance (P)
Pulse P erformance-Alt (P)
Pulse Mo dulation Video
HP/Agilent 3325A/B AM Accuracy (P)
AM Dynamic Range (P)
Maxim um FM Deviation (P)
Mo dulation Meter (P)
Maxim um FM Deviation (P)
P o wer Meter P o wer Range: 1 W to 100 mW
P o wer Accuracy (P)
AM Dynamic Range (P)
ALC P o wer Level
P o wer Meter P o wer Range: 1 W to 100 mW
HP/Agilent 437B P o wer Flatness (P)
P o wer Flatness (A)
P o wer Meter P o wer Range: 1 W to 100 mW
HP/Agilent 438A P o wer Flatness and
P o wer Flatness (AA)
Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.

T able 1-3. Required Equipment for 83630B (continued)
Instrument Critical Sp eci cations Recommended
P o wer Sensor F requency Range: 10 MHz to 2.3 GHz
HP/Agilen t 8482A P o wer Flatness (P)
P o wer Flatness (A)
Step A ttenuator Flatness (AP)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
P o wer Sensor F requency Range: 50 MHz to 26.5 GHz
HP/Agilen t 8485D Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
P o wer Sensor F requency Range: 50 MHz to 26.5 GHz
HP/Agilen t 8485A P o wer Accuracy (P)
P o wer Flatness (P)
AM Dynamic Range (P)
ALC P o wer Level
P o wer Flatness (A)
P o wer Flatness and
P o wer Flatness (AA)
Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
A ttenuator A ttenuation: 30 dB 0.05 dB at 50 MHz HP/Agilen t 11708A Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
F requency Range (tuned):
2.5 MHz to 1.3 GHz
F requency Range (carrier):
HP/Agilen t 3048A Single Sideband Phase
0.01 to 18 GHz
Sensitivit y:
< 70 dBc at 100 Hz o set
< 78 dBc at 1 kHz o set
< 86 dBc at 10 kHz o set
< 107 dBc at 100 kHz o set
O set F requency Range:
100 Hz to 2 MHz
Amplitude Accuracy:
2 dB to 1 MHz o set
1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.
Agilent 8360 Required Equipment

T able 1-3. Required Equipment for 83630B (continued)
Instrument Critical Sp eci cations Recommended
10 MHz Standard (A)
HP/Agilent 8348B FM Bandwidth (P)
HP 8447F Pulse P erformance-Alt (P)
Pulse Mo dulation Video
P o wer Ampli er Gain: 22 dB
1 meter of coax or
P/N 0955-0307 Pulse P erformance-Alt (P)
P o wer Splitter F requency Range: 10 MHz to 26.5 GHz HP/Agilent 11667B FM Bandwidth (P)
HP/Agilent 33330D External Leveling (P)
Pulse P erformance-Alt (P)
Maxim um Input P o wer: 300 mW
Pulse Mo dulation Video
P o wer Accuracy (P)
Maxim um Input P o wer: 300 mW
Pulse P erformance (P)
Pulse P erformance-Alt (P)
FM Bandwidth (P) (2)
Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AA)
1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est
P { Manual P erformance T est

T able 1-3. Required Equipment for 83630B (continued)
Instrument Critical Sp eci cations Recommended
Maxim um Leveled P o wer (P)
Maxim um Input P o wer: 300 mW
Pulse P erformance (P)
3.7 GHz Lo w
P/N 9135-0191 Pulse Mo dulation Video
K & L Microwa v e
Pulse Mo dulation Video
500 MHz Lo w
Pulse Mo dulation Video
T o ol Kit No Substitute P/N 08360{60060
In v ertron California
Capacitor 1000 pf P/N 0160-4574 External Leveling (P)
HP/Agilen t 11975A Spurious Signals
Pulse P erformance-Alt (P)
Pulse P erformance-Alt (P)
WR42 to APC 3.5 mm (f ) K281C Spurious Signals
Pulse P erformance-Alt (P)
1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est
P { Manual P erformance T est
Agilent 8360 Required Equipment

The following list of adapters and cables is provided for con v enience. They may b e used in
equipmen t setups for p erformance tests or adjustmen ts.
SMA (m) to SMA (m) adapter 1250-1159
SMA (f ) to SMA (f ) adapter 1250-1158
SMB (m) to SMB (m) adapter 1250-0669
SMB (f ) to SMB (f ) adapter 1250-0672
SMB tee (f ) (m) (m) 1250-1391
3.5 mm (f ) to 3.5 mm (f ) adapter 5061-5311
3.5 mm (f ) to N-t yp e (m) adapter 1250-1745
2.4 mm (f ) to 2.92 mm (f ) adapter 1250-2187
2.4 mm (f ) to 2.4 mm (f ) adapter 1250-2188
2.4 mm (m) to 3.5 mm (f ) adapter 11901C
BNC (f ) to BNC (f ) adapter 1250-0080
BNC (m) to BNC (m) adapter 1250-0216
BNC (f ) to SMA (m) adapter 1250-1200
BNC (f ) to SMB (m) adapter 1250-1237
BNC tee (m) (f ) (f ) 1250-0781
SMB (f ) to BNC (m) exible cable 85680-60093
SMA semi-rigid cable 2 feet 08340-20124
GPIB cable 2 meters HP/Agilent 10833B
BNC cable 8120-2582 · 1-22 Required Equipment · Agilent 83630B

Agilent 83640B Required Equipment
T able 1-4. Required Equipment for 83640B
Instrument Critical Sp eci cations Recommended
(40 GHz with external mixers)
80 dB display (1.0 dB maxim um)
Pulse Mo dulation On/O
to v ertical p osition of trace on display
Capable of phase lo cking to external
Pulse P erformance-Alt (P)
AM Dynamic Range (P)
Maxim um FM Deviation (P)
and API Spurs (A)
Square W a v e Symmetry (A)
A vs B Sweep Mo de
Pulse P erformance-Alt (P)
Controller 4 Mb yte RAM
Step A ttenuator Flatness (AP)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.
Agilent 8360 Required Equipment

T able 1-4. Required Equipment for 83640B (continued)
Instrument Critical Sp eci cations Recommended
P/N 08360-10001 Step A ttenuator Flatness (AP)
(Shipp ed with instrument)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
AM Dynamic Range (P)
Mo dulation Meter (P)
Lo w P o wer SRD Bias (A)
F requency Switching Time (P)
Pulse P erformance-Alt (P)
Pulse Mo dulation Video
In ternal Pulse Accuracy (P)
10 MHz Standard (A)
Mo dulator O set and Gain (A)
Pulse P erformance (P)
scale 2 mV
T rigger Sensitivit y DC to 100 MHz:
T rigger & Timebase Jitter:
2.5 ps + 5E
Time In terv al Accuracy:
10 ps 0.1% of reading
Channel Input Re ection:
5% for 30 ps risetime
1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.

T able 1-4. Required Equipment for 83640B (continued)
Instrument Critical Sp eci cations Recommended
F requency Switc hing Time (P)
Pulse P erformance-Alt (P)
Pulse Mo dulation Video
Mo dulator O set and Gain (A)
Pulse P erformance (P)
Pulse P erformance-Alt (P)
Pulse Mo dulation Video
HP/Agilen t 3325A/B AM Accuracy (P)
AM Dynamic Range (P)
Maxim um FM Deviation (P)
Mo dulation Meter (P)
Maxim um FM Deviation (P)
P o wer Meter P o wer Range: 1 W to 100 mW
P o wer Accuracy (P)
AM Dynamic Range (P)
ALC P o wer Level
P o wer Meter P o wer Range: 1 W to 100 mW
HP/Agilen t 437B P o wer Flatness (P)
P o wer Flatness (A)
P o wer Meter P o wer Range: 1 W to 100 mW
HP/Agilen t 438A P o wer Flatness and
Step A ttenuator Flatness (AP)
P o wer Flatness (AA)
Step A ttenuator Flatness (AA)
1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.
Agilent 8360 Required Equipment

T able 1-4. Required Equipment for 83640B (continued)
Instrument Critical Sp eci cations Recommended
P o wer Sensor F requency Range: 10 MHz to 2.3 GHz
HP/Agilent 8482A P o wer Flatness (P)
P o wer Flatness (A)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
Step A ttenuator Flatness (AP)
P o wer Sensor F requency Range: 50 MHz to 40 GHz
HP/Agilent 8487A P o wer Accuracy (P)
P o wer Flatness (P)
AM Dynamic Range (P)
ALC P o wer Level
P o wer Flatness (A)
Square W a v e Symmetry (A)
Step A ttenuator Flatness (AP)
P o wer Flatness (AA)
Step A ttenuator Flatness (AA)
P o wer Sensor F requency Range: 50 MHz to 40 GHz
HP/Agilent 8487D Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
A ttenuator A ttenuation: 30 dB 0.05 dB at 50 MHz HP/Agilent 11708A Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
F requency Range (tuned):
2.5 MHz to 1.3 GHz
Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
F requency Range (carrier):
HP/Agilent 3048A Single Sideband Phase
0.01 to 18 GHz
Sensitivit y:
< 70 dBc at 100 Hz o set
< 78 dBc at 1 kHz o set
< 86 dBc at 10 kHz o set
< 107 dBc at 100 kHz o set
O set F requency Range:
100 Hz to 2 MHz
Amplitude Accuracy:
2 dB to 1 MHz o set
1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.

T able 1-4. Required Equipment for 83640B (continued)
Instrument Critical Sp eci cations Recommended
10 MHz Standard (A)
HP/Agilen t 8348A FM Bandwidth (P)
HP/Agilen t 8447F Pulse P erformance-Alt (P)
Pulse Mo dulation Video
P o wer Ampli er Gain: 22 dB
1 meter of coax or
P/N 0955-0307 Pulse P erformance-Alt (P)
P o wer Splitter F requency Range: 10 MHz to 26.5 GHz HP/Agilen t 11667C FM Bandwidth (P)
HP/Agilen t 33330E External Leveling (P)
Pulse P erformance-Alt (P)
Maxim um Input P o wer: 300 mW
Pulse Mo dulation Video
Maxim um Leveled P o wer (P)
Maxim um Input P o wer: 300 mW
Pulse P erformance (P)
P o wer Accuracy (P)
Maxim um Input P o wer: 300 mW
Pulse P erformance (P)
Pulse P erformance-Alt (P)
FM Bandwidth (P) (2)
Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AA)
1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est
P { Manual P erformance T est
Agilent 8360 Required Equipment

T able 1-4. Required Equipment for 83640B (continued)
Instrument Critical Sp eci cations Recommended
3.7 GHz Lo w
P/N 9135-0191 Pulse Mo dulation Video
K & L Microwa v e
Pulse Mo dulation Video
500 MHz Lo w
Pulse Mo dulation Video
T o ol Kit No Substitute P/N 08360{60060
In v ertron California
Capacitor 1000 pf P/N 0160-4574 External Leveling (P)
HP/Agilent 11975A Spurious Signals
Pulse P erformance-Alt (P)
Pulse P erformance-Alt (P)
Pulse P erformance-Alt (P)
WR42 to APC 3.5 mm (f ) K281C Spurious Signals
Pulse P erformance-Alt (P)
WR28 to APC 2.4 mm (f ) R281A Spurious Signals
Pulse P erformance-Alt (P)
1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est
P { Manual P erformance T est

The follo wing list of adapters and cables is provided for con v enience. They may b e used in
equipmen t setups for p erformance tests or adjustmen ts.
SMA (m) to SMA (m) adapter 1250-1159
SMA (f ) to SMA (f ) adapter 1250-1158
SMB (m) to SMB (m) adapter 1250-0669
SMB (f ) to SMB (f ) adapter 1250-0672
SMB tee (f ) (m) (m) 1250-1391
3.5 mm (f ) to 3.5 mm (f ) adapter 5061-5311
3.5 mm (f ) to N-t yp e (m) adapter 1250-1745
2.4 mm (f ) to 2.92 mm (f ) adapter 1250-2187
2.4 mm (f ) to 2.4 mm (f ) adapter 1250-2188
2.4 mm (m) to 3.5 mm (f ) adapter 11901C
BNC (f ) to BNC (f ) adapter 1250-0080
BNC (m) to BNC (m) adapter 1250-0216
BNC (f ) to SMA (m) adapter 1250-1200
BNC (f ) to SMB (m) adapter 1250-1237
BNC tee (m) (f ) (f ) 1250-0781
SMB (f ) to BNC (m) exible cable 85680-60093
SMA semi-rigid cable 2 feet 08340-20124
GPIB cable 2 meters HP/Agilent 10833B
Agilent 8360 Required Equipment
BNC cable 8120-2582 · Agilent 83640B

Agilent 83650B Required Equipment
T able 1-5. Required Equipment for 83650B
Instrument Critical Sp eci cations Recommended
(50 GHz with external mixers)
80 dB display (1.0 dB maxim um)
Pulse Mo dulation On/O
to v ertical p osition of trace on display
Capable of phase lo cking to external
Pulse P erformance-Alt (P)
AM Dynamic Range (P)
Maxim um FM Deviation (P)
and API Spurs (A)
Square W a v e Symmetry (A)
A vs B Sweep Mo de
Pulse P erformance-Alt (P)
Controller 4 Mb yte RAM
Step A ttenuator Flatness (AP)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.

T able 1-5. Required Equipment for 83650B (continued)
Instrument Critical Sp eci cations Recommended
P/N 08360-10001 Step A ttenuator Flatness (AP)
(Shipp ed with instrument)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
AM Dynamic Range (P)
Mo dulation Meter (P)
Lo w P o wer SRD Bias (A)
F requency Switc hing Time (P)
Pulse P erformance-Alt (P)
Pulse Mo dulation Video
In ternal Pulse Accuracy (P)
10 MHz Standard (A)
Mo dulator O set and
Pulse P erformance (P)
scale 2 mV
T rigger Sensitivit y DC to 100 MHz:
T rigger & Timebase Jitter:
2.5 ps + 5E
Time In terv al Accuracy:
10 ps 0.1% of reading
Channel Input Re ection:
5% for 30 ps risetime
1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.
Agilent 8360 Required Equipment

T able 1-5. Required Equipment for 83650B (continued)
Instrument Critical Sp eci cations Recommended
F requency Switching Time (P)
Pulse P erformance-Alt (P)
Pulse Mo dulation Video
Mo dulator O set and
Pulse P erformance (P)
Pulse P erformance-Alt (P)
Pulse Mo dulation Video
HP/Agilent 3325A/B AM Accuracy (P)
AM Dynamic Range (P)
Maxim um FM Deviation (P)
Mo dulation Meter (P)
Maxim um FM Deviation (P)
P o wer Meter P o wer Range: 1 W to 100 mW
P o wer Accuracy (P)
AM Dynamic Range (P)
ALC P o wer Level
P o wer Meter P o wer Range: 1 W to 100 mW
HP/Agilent 437B P o wer Flatness (P)
P o wer Flatness (A)
P o wer Meter P o wer Range: 1 W to 100 mW
HP/Agilent 438A P o wer Flatness and
Step A ttenuator Flatness (AP)
P o wer Flatness (AA)
Step A ttenuator Flatness (AA)
1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.

T able 1-5. Required Equipment for 83650B (continued)
Instrument Critical Sp eci cations Recommended
P o wer Sensor F requency Range: 10 MHz to 2.3 GHz
HP/Agilen t 8482A P o wer Flatness (P)
P o wer Flatness (A)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
Step A ttenuator Flatness (AP)
P o wer Sensor F requency Range: 50 MHz to 50 GHz
HP/Agilen t 8487A P o wer Accuracy (P)
P o wer Flatness (P)
AM Dynamic Range (P)
ALC P o wer Level
P o wer Flatness (A)
P o wer Flatness and
Step A ttenuator Flatness (AP)
P o wer Flatness (AA)
Step A ttenuator Flatness (AA)
P o wer Sensor F requency Range: 50 MHz to 50 GHz
HP/Agilen t 8487D Step A ttenuator Flatness (AP)
Step A ttenuator Flatness (AA)
A ttenuator A ttenuation: 30 dB 0.05 dB at 50 MHz HP/Agilen t 11708A Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
F requency Range (tuned):
2.5 MHz to 1.3 GHz
Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
F requency Range (carrier):
HP/Agilen t 3048A Single Sideband Phase
0.01 to 18 GHz
Sensitivit y:
< 70 dBc at 100 Hz o set
< 78 dBc at 1 kHz o set
< 86 dBc at 10 kHz o set
< 107 dBc at 100 kHz o set
O set F requency Range:
100 Hz to 2 MHz
Amplitude Accuracy:
2 dB to 1 MHz o set
1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.
Agilent 8360 Required Equipment

T able 1-5. Required Equipment for 83650B (continued)
Instrument Critical Sp eci cations Recommended
10 MHz Standard (A)
HP/Agilent 8348A FM Bandwidth (P)
HP/Agilent 8447F Pulse P erformance-Alt (P)
Pulse Mo dulation Video
P o wer Ampli er Gain: 22 dB
1 meter of coax or
P/N 0955-0307 Pulse P erformance-Alt (P)
P o wer Splitter F requency Range: 10 MHz to 50 GHz HP/Agilent 11667C FM Bandwidth (P)
HP/Agilent 33330E External Leveling (P)
Pulse P erformance-Alt (P)
Maxim um Input P o wer: 300 mW
Pulse Mo dulation Video
Maxim um Leveled P o wer (P)
Maxim um Input P o wer: 300 mW
Pulse P erformance (P)
P o wer Accuracy (P)
Maxim um Input P o wer: 300 mW
Pulse P erformance (P)
Pulse P erformance-Alt (P)
FM Bandwidth (P) (2)
Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AA)
1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est
P { Manual P erformance T est

T able 1-5. Required Equipment for 83650B (continued)
Instrument Critical Sp eci cations Recommended
3.7 GHz Lo w
P/N 9135-0191 Pulse Mo dulation Video
K & L Microwa v e
Pulse Mo dulation Video
500 MHz Lo w
Pulse Mo dulation Video
T o ol Kit No Substitute P/N 08360{60060
In v ertron California
Capacitor 1000 pf P/N 0160-4574 External Leveling (P)
Pulse P erformance-Alt (P)
HP/Agilen t 11975A Spurious Signals
Pulse P erformance-Alt (P)
Pulse P erformance-Alt (P)
Pulse P erformance-Alt (P)
WR42 to APC 3.5 mm (f ) K281C Spurious Signals
Pulse P erformance-Alt (P)
1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est
P { Manual P erformance T est
Agilent 8360 Required Equipment

T able 1-5. Required Equipment for 83650B (continued)
Instrument Critical Sp eci cations Recommended
WR28 to APC 2.4 mm (f ) R281A Spurious Signals
Pulse P erformance-Alt (P)
WR22 to APC 2.4 mm (f ) Q281A Spurious Signals
Pulse P erformance-Alt (P)
1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est
P { Manual P erformance T est

The follo wing list of adapters and cables is provided for con v enience. They may b e used in
equipmen t setups for p erformance tests or adjustmen ts.
SMA (m) to SMA (m) adapter 1250-1159
SMA (f ) to SMA (f ) adapter 1250-1158
SMB (m) to SMB (m) adapter 1250-0669
SMB (f ) to SMB (f ) adapter 1250-0672
SMB tee (f ) (m) (m) 1250-1391
3.5 mm (f ) to 3.5 mm (f ) adapter 5061-5311
3.5 mm (f ) to N-t yp e (m) adapter 1250-1745
2.4 mm (f ) to 2.92 mm (f ) adapter 1250-2187
2.4 mm (f ) to 2.4 mm (f ) adapter 1250-2188
2.4 mm (m) to 3.5 mm (f ) adapter 11901C
BNC (f ) to BNC (f ) adapter 1250-0080
BNC (m) to BNC (m) adapter 1250-0216
BNC (f ) to SMA (m) adapter 1250-1200
BNC (f ) to SMB (m) adapter 1250-1237
BNC tee (m) (f ) (f ) 1250-0781
SMB (f ) to BNC (m) exible cable 85680-60093
SMA semi-rigid cable 2 feet 08340-20124
GPIB cable 2 meters HP/Agilent 10833B
Agilent 8360 Required Equipment
BNC cable 8120-2582 · Agilent 83650B

Agilent 83623L Required Equipment
T able 1-6. Required Equipment for 83623L
Instrument Critical Sp eci cations Recommended
80 dB display (1.0 dB maxim um)
to v ertical p osition of trace on display
Capable of phase lo cking to external
and API Spurs (A)
Square W a v e Symmetry (A)
A vs B Sweep Mo de
AM O set (A)
Controller 4 Mb yte RAM
Step A ttenuator Flatness (AP)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
P/N 08360-10001 Step A ttenuator Flatness (AP)
(Shipp ed with instrument)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
Lo w P o wer SRD Bias (A)
1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.

T able 1-6. Required Equipment for 83623L (continued)
Instrument Critical Sp eci cations Recommended
F requency Switc hing Time (P)
10 MHz Standard (A)
F requency Switc hing Time (P)
P o wer Meter P o wer Range: 1 W to 100 mW
P o wer Accuracy (P)
ALC P o wer Level
P o wer Meter P o wer Range: 1 W to 100 mW
HP/Agilen t 437B P o wer Flatness (P)
P o wer Flatness (A)
P o wer Meter P o wer Range: 1 W to 100 mW
HP/Agilen t 438A P o wer Flatness and
P o wer Flatness (AA)
Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.
Agilent 8360 Required Equipment

T able 1-6. Required Equipment for 83623L (continued)
Instrument Critical Sp eci cations Recommended
P o wer Sensor F requency Range: 10 MHz to 2.3 GHz
HP/Agilent 8482A P o wer Flatness (P)
P o wer Flatness (A)
Step A ttenuator Flatness (AP)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
P o wer Sensor F requency Range: 50 MHz to 20 GHz
HP/Agilent 8485D Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
P o wer Sensor F requency Range: 50 MHz to 20 GHz
HP/Agilent 8485A P o wer Accuracy (P)
P o wer Flatness (P)
ALC P o wer Level
P o wer Flatness (A)
P o wer Flatness and
P o wer Flatness (AA)
Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
A ttenuator A ttenuation: 30 dB 0.05 dB at 50 MHz HP/Agilent 11708A Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
F requency Range (carrier):
HP/Agilent 3048A Single Sideband Phase
0.01 to 18 GHz
Sensitivit y:
< 70 dBc at 100 Hz o set
< 78 dBc at 1 kHz o set
< 86 dBc at 10 kHz o set
< 107 dBc at 100 kHz o set
O set F requency Range:
100 Hz to 2 MHz
Amplitude Accuracy:
2 dB to 1 MHz o set
10 MHz Standard (A)
HP/Agilent 33330D External Leveling (P)
AM O set (A)
1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est
P { Manual P erformance T est

T able 1-6. Required Equipment for 83623L (continued)
Instrument Critical Sp eci cations Recommended
P o wer Accuracy (P)
Maxim um Input P o wer: 300 mW
Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AA)
Maxim um Leveled P o wer (P)
Maxim um Input P o wer: 300 mW
T o ol Kit No Substitute P/N 08360{60060
In v ertron California
Capacitor 1000 pf P/N 0160-4574 External Leveling (P)
1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est
P { Manual P erformance T est
Agilent 8360 Required Equipment

The following list of adapters and cables is provided for con v enience. They may b e used in
equipmen t setups for p erformance tests or adjustmen ts.
SMA (m) to SMA (m) adapter 1250-1159
SMA (f ) to SMA (f ) adapter 1250-1158
SMB (m) to SMB (m) adapter 1250-0669
SMB (f ) to SMB (f ) adapter 1250-0672
SMB tee (f ) (m) (m) 1250-1391
3.5 mm (f ) to 3.5 mm (f ) adapter 5061-5311
3.5 mm (f ) to N-t yp e (m) adapter 1250-1745
2.4 mm (f ) to 2.92 mm (f ) adapter 1250-2187
2.4 mm (f ) to 2.4 mm (f ) adapter 1250-2188
2.4 mm (m) to 3.5 mm (f ) adapter 11901C
BNC (f ) to BNC (f ) adapter 1250-0080
BNC (m) to BNC (m) adapter 1250-0216
BNC (f ) to SMA (m) adapter 1250-1200
BNC (f ) to SMB (m) adapter 1250-1237
BNC tee (m) (f ) (f ) 1250-0781
SMB (f ) to BNC (m) exible cable 85680-60093
SMA semi-rigid cable 2 feet 08340-20124
GPIB cable 2 meters HP/Agilent 10833B
BNC cable 8120-2582 · 1-42 Required Equipment · Agilent 83623L

Agilent 83630L Required Equipment
T able 1-7. Required Equipment for 83630L
Instrument Critical Sp eci cations Recommended
(26.5 GHz with external mixers)
80 dB display (1.0 dB maxim um)
to v ertical p osition of trace on display
and API Spurs (A)
Capable of phase lo cking to external
Square W a v e Symmetry (A)
A vs B Sweep Mo de
Controller 4 Mb yte RAM
Step A ttenuator Flatness (AP)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
P/N 08360-10001 Step A ttenuator Flatness (AP)
(Shipp ed with instrument)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
Lo w P o wer SRD Bias (A)
1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.
Agilent 8360 Required Equipment

T able 1-7. Required Equipment for 83630L (continued)
Instrument Critical Sp eci cations Recommended
F requency Switching Time (P)
10 MHz Standard (A)
F requency Switching Time (P)
AM O set (A)
P o wer Meter P o wer Range: 1 W to 100 mW
P o wer Accuracy (P)
ALC P o wer Level
P o wer Meter P o wer Range: 1 W to 100 mW
HP/Agilent 437B P o wer Flatness (P)
P o wer Flatness (A)
P o wer Meter P o wer Range: 1 W to 100 mW
HP/Agilent 438A P o wer Flatness and
P o wer Flatness (AA)
Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
P o wer Sensor F requency Range: 10 MHz to 2.3 GHz
HP/Agilent 8482A P o wer Flatness (P)
P o wer Flatness (A)
Step A ttenuator Flatness (AP)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.

T able 1-7. Required Equipment for 83630L (continued)
Instrument Critical Sp eci cations Recommended
P o wer Sensor F requency Range: 50 MHz to 26.5 GHz
HP/Agilen t 8485D Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
P o wer Sensor F requency Range: 50 MHz to 26.5 GHz
HP/Agilen t 8485A P o wer Accuracy (P)
P o wer Flatness (P)
ALC P o wer Level
P o wer Flatness (A)
P o wer Flatness and
P o wer Flatness (AA)
Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
A ttenuator A ttenuation: 30 dB 0.05 dB at 50 MHz HP/Agilen t 11708A Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
F requency Range (carrier):
HP/Agilen t 3048A Single Sideband Phase
0.01 to 18 GHz
Sensitivit y:
< 70 dBc at 100 Hz o set
< 78 dBc at 1 kHz o set
< 86 dBc at 10 kHz o set
< 107 dBc at 100 kHz o set
O set F requency Range:
100 Hz to 2 MHz
Amplitude Accuracy:
2 dB to 1 MHz o set
10 MHz Standard (A)
HP/Agilen t 33330D External Leveling (P)
AM O set (A)
1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est
P { Manual P erformance T est
Agilent 8360 Required Equipment

T able 1-7. Required Equipment for 83630L (continued)
Instrument Critical Sp eci cations Recommended
P o wer Accuracy (P)
Maxim um Input P o wer: 300 mW
Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AA)
Maxim um Leveled P o wer (P)
Maxim um Input P o wer: 300 mW
T o ol Kit No Substitute P/N 08360{60060
In v ertron California
Capacitor 1000 pf P/N 0160-4574 External Leveling (P)
HP/Agilent 11975A Spurious Signals
WR42 to APC 3.5 mm (f ) K281C Spurious Signals
1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est
P { Manual P erformance T est

The follo wing list of adapters and cables is provided for con v enience. They may b e used in
equipmen t setups for p erformance tests or adjustmen ts.
SMA (m) to SMA (m) adapter 1250-1159
SMA (f ) to SMA (f ) adapter 1250-1158
SMB (m) to SMB (m) adapter 1250-0669
SMB (f ) to SMB (f ) adapter 1250-0672
SMB tee (f ) (m) (m) 1250-1391
3.5 mm (f ) to 3.5 mm (f ) adapter 5061-5311
3.5 mm (f ) to N-t yp e (m) adapter 1250-1745
2.4 mm (f ) to 2.92 mm (f ) adapter 1250-2187
2.4 mm (f ) to 2.4 mm (f ) adapter 1250-2188
2.4 mm (m) to 3.5 mm (f ) adapter 11901C
BNC (f ) to BNC (f ) adapter 1250-0080
BNC (m) to BNC (m) adapter 1250-0216
BNC (f ) to SMA (m) adapter 1250-1200
BNC (f ) to SMB (m) adapter 1250-1237
BNC tee (m) (f ) (f ) 1250-0781
SMB (f ) to BNC (m) exible cable 85680-60093
SMA semi-rigid cable 2 feet 08340-20124
GPIB cable 2 meters HP/Agilent 10833B
Agilent 8360 Required Equipment
BNC cable 8120-2582 · Agilent 83630L

Agilent 83640L Required Equipment
T able 1-8. Required Equipment for 83640L
Instrument Critical Sp eci cations Recommended
(40 GHz with external mixers)
80 dB display (1.0 dB maxim um)
to v ertical p osition of trace on display
and API Spurs (A)
Capable of phase lo cking to external
Square W a v e Symmetry (A)
A vs B Sweep Mo de
AM O set (A)
Controller 4 Mb yte RAM
Step A ttenuator Flatness (AP)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
P/N 08360-10001 Step A ttenuator Flatness (AP)
(Shipp ed with instrument)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
Lo w P o wer SRD Bias (A)
1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.

T able 1-8. Required Equipment for 83640L (continued)
Instrument Critical Sp eci cations Recommended
F requency Switc hing Time (P)
10 MHz Standard (A)
F requency Switc hing Time (P)
P o wer Meter P o wer Range: 1 W to 100 mW
P o wer Accuracy (P)
ALC P o wer Level
P o wer Meter P o wer Range: 1 W to 100 mW
HP/Agilen t 437B P o wer Flatness (P)
P o wer Flatness (A)
P o wer Meter P o wer Range: 1 W to 100 mW
HP/Agilen t 438A P o wer Flatness and
Step A ttenuator Flatness (AP)
P o wer Flatness (AA)
Step A ttenuator Flatness (AA)
P o wer Sensor F requency Range: 10 MHz to 2.3 GHz
HP/Agilen t 8482A P o wer Flatness (P)
P o wer Flatness (A)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
Step A ttenuator Flatness (AP)
1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.
Agilent 8360 Required Equipment

T able 1-8. Required Equipment for 83640L (continued)
Instrument Critical Sp eci cations Recommended
P o wer Sensor F requency Range: 50 MHz to 40 GHz
HP/Agilent 8487A P o wer Accuracy (P)
P o wer Flatness (P)
ALC P o wer Level
P o wer Flatness (A)
Square W a v e Symmetry (A)
Step A ttenuator Flatness (AP)
P o wer Flatness (AA)
Step A ttenuator Flatness (AA)
P o wer Sensor F requency Range: 50 MHz to 40 GHz
HP/Agilent 8487D Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
A ttenuator A ttenuation: 30 dB 0.05 dB at 50 MHz HP/Agilent 11708A Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
F requency Range (tuned):
Step A ttenuator Flatness (AA)
2.5 MHz to 1.3 GHz
Step A ttenuator Flatness (AP)
F requency Range (carrier):
HP/Agilent 3048A Single Sideband Phase
0.01 to 18 GHz
Sensitivit y:
< 70 dBc at 100 Hz o set
< 78 dBc at 1 kHz o set
< 86 dBc at 10 kHz o set
< 107 dBc at 100 kHz o set
O set F requency Range:
100 Hz to 2 MHz
Amplitude Accuracy:
2 dB to 1 MHz o set
10 MHz Standard (A)
HP/Agilent 33330E External Leveling (P)
AM O set (A)
1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.

T able 1-8. Required Equipment for 83640L (continued)
Instrument Critical Sp eci cations Recommended
Maxim um Leveled P o wer (P)
Maxim um Input P o wer: 300 mW
P o wer Accuracy (P)
Maxim um Input P o wer: 300 mW
Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AA)
T o ol Kit No Substitute P/N 08360{60060
In v ertron California
Capacitor 1000 pf P/N 0160-4574 External Leveling (P)
HP/Agilen t 11975A Spurious Signals
WR42 to APC 3.5 mm (f ) K281C Spurious Signals
WR28 to APC 2.4 mm (f ) R281A Spurious Signals
1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est
P { Manual P erformance T est
Agilent 8360 Required Equipment

The following list of adapters and cables is provided for con v enience. They may b e used in
equipmen t setups for p erformance tests or adjustmen ts.
SMA (m) to SMA (m) adapter 1250-1159
SMA (f ) to SMA (f ) adapter 1250-1158
SMB (m) to SMB (m) adapter 1250-0669
SMB (f ) to SMB (f ) adapter 1250-0672
SMB tee (f ) (m) (m) 1250-1391
3.5 mm (f ) to 3.5 mm (f ) adapter 5061-5311
3.5 mm (f ) to N-t yp e (m) adapter 1250-1745
2.4 mm (f ) to 2.92 mm (f ) adapter 1250-2187
2.4 mm (f ) to 2.4 mm (f ) adapter 1250-2188
2.4 mm (m) to 3.5 mm (f ) adapter 11901C
BNC (f ) to BNC (f ) adapter 1250-0080
BNC (m) to BNC (m) adapter 1250-0216
BNC (f ) to SMA (m) adapter 1250-1200
BNC (f ) to SMB (m) adapter 1250-1237
BNC tee (m) (f ) (f ) 1250-0781
SMB (f ) to BNC (m) exible cable 85680-60093
SMA semi-rigid cable 2 feet 08340-20124
GPIB cable 2 meters HP/Agilent 10833B
BNC cable 8120-2582 · 1-52 Required Equipment · Agilent 83640L

Agilent 83650L Required Equipment
T able 1-9. Required Equipment for 83650L
Instrument Critical Sp eci cations Recommended
(50 GHz with external mixers)
80 dB display (1.0 dB maxim um)
to v ertical p osition of trace on display
and API Spurs (A)
Capable of phase lo cking to external
Square W a v e Symmetry (A)
A vs B Sweep Mo de
AM O set (A)
Controller 4 Mb yte RAM
Step A ttenuator Flatness (AP)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
P/N 08360-10001 Step A ttenuator Flatness (AP)
(Shipp ed with instrument)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
Lo w P o wer SRD Bias (A)
1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.
Agilent 8360 Required Equipment

T able 1-9. Required Equipment for 83650L (continued)
Instrument Critical Sp eci cations Recommended
F requency Switching Time (P)
10 MHz Standard (A)
F requency Switching Time (P)
P o wer Meter P o wer Range: 1 W to 100 mW
P o wer Accuracy (P)
ALC P o wer Level
P o wer Meter P o wer Range: 1 W to 100 mW
HP/Agilent 437B P o wer Flatness (P)
P o wer Flatness (A)
P o wer Meter P o wer Range: 1 W to 100 mW
HP/Agilent 438A P o wer Flatness and
Step A ttenuator Flatness (AP)
P o wer Flatness (AA)
Step A ttenuator Flatness (AA)
P o wer Sensor F requency Range: 10 MHz to 2.3 GHz
HP/Agilent 8482A P o wer Flatness (P)
P o wer Flatness (A)
P o wer Flatness and
Step A ttenuator Flatness (AA)
P o wer Flatness (AA)
Step A ttenuator Flatness (AP)
1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.

T able 1-9. Required Equipment for 83650L (continued)
Instrument Critical Sp eci cations Recommended
P o wer Sensor F requency Range: 50 MHz to 50 GHz
HP/Agilen t 8487A P o wer Accuracy (P)
P o wer Flatness (P)
ALC P o wer Level
P o wer Flatness (A)
P o wer Flatness and
Step A ttenuator Flatness (AP)
P o wer Flatness (AA)
Step A ttenuator Flatness (AA)
P o wer Sensor F requency Range: 50 MHz to 50 GHz
HP/Agilen t 8487D Step A ttenuator Flatness (AP)
Step A ttenuator Flatness (AA)
A ttenuator A ttenuation: 30 dB 0.05 dB at 50 MHz HP/Agilen t 11708A Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AP)
F requency Range (tuned):
Step A ttenuator Flatness (AA)
2.5 MHz to 1.3 GHz
Step A ttenuator Flatness (AP)
F requency Range (carrier):
HP/Agilen t 3048A Single Sideband Phase
0.01 to 18 GHz
Sensitivit y:
< 70 dBc at 100 Hz o set
< 78 dBc at 1 kHz o set
< 86 dBc at 10 kHz o set
< 107 dBc at 100 kHz o set
O set F requency Range:
100 Hz to 2 MHz
Amplitude Accuracy:
2 dB to 1 MHz o set
10 MHz Standard (A)
HP/Agilen t 33330E External Leveling (P)
AM O set (A)
1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est
P { Manual P erformance T est
2 Recommended mo del is part of the microw ave test station.
Agilent 8360 Required Equipment

T able 1-9. Required Equipment for 83650L (continued)
Instrument Critical Sp eci cations Recommended
Maxim um Leveled P o wer (P)
Maxim um Input P o wer: 300 mW
Pulse P erformance (P)
P o wer Accuracy (P)
Maxim um Input P o wer: 300 mW
Step A ttenuator Flatness (AA)
Step A ttenuator Flatness (AA)
T o ol Kit No Substitute P/N 08360{60060
In v ertron California
Capacitor 1000 pf P/N 0160-4574 External Leveling (P)
HP/Agilent 11975A Spurious Signals
WR42 to APC 3.5 mm (f ) K281C Spurious Signals
WR28 to APC 2.4 mm (f ) R281A Spurious Signals
WR22 to APC 2.4 mm (f ) Q281A Spurious Signals
1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est
P { Manual P erformance T est

The follo wing list of adapters and cables is provided for con v enience. They may b e used in
equipmen t setups for p erformance tests or adjustmen ts.
SMA (m) to SMA (m) adapter 1250-1159
SMA (f ) to SMA (f ) adapter 1250-1158
SMB (m) to SMB (m) adapter 1250-0669
SMB (f ) to SMB (f ) adapter 1250-0672
SMB tee (f ) (m) (m) 1250-1391
3.5 mm (f ) to 3.5 mm (f ) adapter 5061-5311
3.5 mm (f ) to N-t yp e (m) adapter 1250-1745
2.4 mm (f ) to 2.92 mm (f ) adapter 1250-2187
2.4 mm (f ) to 2.4 mm (f ) adapter 1250-2188
2.4 mm (m) to 3.5 mm (f ) adapter 11901C
BNC (f ) to BNC (f ) adapter 1250-0080
BNC (m) to BNC (m) adapter 1250-0216
BNC (f ) to SMA (m) adapter 1250-1200
BNC (f ) to SMB (m) adapter 1250-1237
BNC tee (m) (f ) (f ) 1250-0781
SMB (f ) to BNC (m) exible cable 85680-60093
SMA semi-rigid cable 2 feet 08340-20124
GPIB cable 2 meters HP/Agilent 10833B
Agilent 8360 Required Equipment
BNC cable 8120-2582 · Agilent 83650L



P erformance T ests
How to Use This Chapter
Use the pro cedures in this c hapter to test the electrical p erformance of the syn thesizer. These
tests do not require access to the in terior of the instrumen t.
The syn thesizer m ust w arm up for at least one hour b efore the electrical sp eci cations are
v alid.
F old-out men u maps are lo cated in Chapter 6. Some men us ha v e more than one page of
softkeys. Select the
more softkey to view the next page of softkeys.
more is not included in
the k eystrokes given in these pro cedures.
In all cases where y ou are instructed to preset the syn thesizer, use the factory
preset mo de only .
F or assurance that most of the in ternal functions of the instrumen t w ork (without testing for
sp eci cations), see \Op erator's Chec k," in y our syn thesizer's User's guide.
Agilent 8360 P erformance T ests 2-1
Menu Maps · Operator's Check

List of T ests
The following is a list of this c hapter's tests and the page n um b er that eac h test is lo cated on:
Op eration V eri cation T ests : 2-5
P erformance T ests : 2-8
P erformance T ests for B-Series Only
The p erformance tests that apply to the
Agilent 8360 B-series swept signal generators
only are marked with the following symb ol: y
Operation V erification T ests
T o meet the needs of most incoming insp ections
(80% v eri cation) , and provide reasonable assurance that the instrumen t w orks, see the
\Op eration V eri cation" section of this c hapter.
Automated P erformance T ests
The automated p erformance tests av ailable at time-of-shipment are included in Chapter 5.
The Step A tten uator test is lo cated in Chapter 5.
2-2 P erformance T ests Agilent 8360

The equipmen t required to p erform the tests in this c hapter is listed in Chapter 1. Y ou may
use an y equipmen t that meets the critical sp eci cations given.
T est records are supplied at the end of this c hapter. Use a test record when y ou p erform a full
calibration of y our syn thesizer. This form provides a tabulated index of the p erformance tests,
their acceptable limits, the measuremen t uncertain ty of eac h test, and a column to record
actual measuremen ts.
There are sev eral test records. Be sure y ou use the one designated (at the top) for y our
syn thesizer. Eac h test record is sho wn b elow. The page n um b er of eac h test record is also
included for easy access.
T est Record for Agilent 83620B /22B : 2-139
T est Record for Agilent 83623B /24B : 2-147
T est Record for Agilent 83630B : 2-155
T est Record for Agilent 83640B : 2-163
T est Record for Agilent 83650B : 2-171
T est Record for Agilent 83623L : 2-179
T est Record for Agilent 83630L : 2-184
T est Record for Agilent 83640L : 2-189
T est Record for Agilent 83650L : 2-194
P erform the tests in the order that they app ear.
P erform the tests in this c hapter at least once ev ery 24 mon ths.
Agilent 8360 P erformance T ests 2-3

The following pro cedures meet the needs of most incoming insp ections (80% v eri cation), and
provide a reasonable assurance that the instrumen t is functioning prop erly . The Op eration
V eri cation F orm is lo cated b elow.
P erform these op eration v eri cation pro cedures in the order given and record the pass/fail
results in the Op eration V eri cation F orm.
T able 2-1. Operation V erification Form
T est F acilit y Date
Serial Numb er T ested By
Options Firmw are Revision
T est Results (P ass/F ail)
2-4 P erformance T ests
Operation V erification · Mo del Customer · Operation V erification

equipmen t is required. Press
Selftest (Full) .
The syn thesizer executes a series of self-tests. When completed, the follo wing message is
display ed if all the tests passed:
Occasionally , the self-tests may fail when it is run remotely or when external
cables are connected to the instrumen t. If the self-tests fail, remov e all cables
and run the self-tests from the front panel. If the self-tests pass under these
conditions, the instrumen t is op erating prop erly .
External connections to the instrumen t during self-test may not allo w the lines
to c hange conditions within the limits of the self-test's parameters.
Accuracy ."
If the syn thesizer fails one or more self-tests, a di eren t message is display ed listing the
most indep enden t test that failed and a reference to the b est en try p oint in to the A gilent
T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series Swept CW Gener ator
T r oublesho oting Guide . Record F ail on the op eration v eri cation form and see the A gilent
T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series Swept CW Gener ator
T r oublesho oting Guide . After repair, rep eat op eration v eri cation from the b eginning.
con tinue to the next test, \Pow er Flatness."
If the syn thesizer fails, record F ail on the op eration v eri cation form and then follo w the
\In Case of Diculty" instructions given for this p erformance test. After repair, rep eat
op eration v eri cation from the b eginning.
Agilent 8360 P erformance T ests
· Operation V erification

con tinue to the next test, \Maxim um Lev eled P ow er."
If the syn thesizer fails, record F ail on the op eration v eri cation form and then follow the
\In Case of Diculty" instructions given for this p erformance test. After repair, rep eat
op eration v eri cation from the b eginning.
Op eration v eri cation is completed.
If the syn thesizer fails, record F ail on the op eration v eri cation form and then follow the
\In Case of Diculty" instructions given for this p erformance test. After repair, rep eat
op eration v eri cation from the b eginning.
2-6 P erformance T ests
Operation V erification

This pro cedure c hec ks the accuracy of the in ternal timebase. The time required for a sp eci c
phase c hange is measured b oth b efore and after a sp eci ed w aiting p erio d. The aging rate is
in v ersely prop ortional to the absolute v alue of the di erence in the measured times.
The ov erall accuracy of the in ternal timebase is a function of:
TBC AR TE LE where:
TBC = timebase calibration
TE = temperature effects
AR = aging rate
LE = line effects
After the timebase is adjusted, the timebase frequency should stay within the aging rate if the
follo wing things happ en:
The timebase ov en do es not co ol do wn.
The instrumen t k eeps the same orientation with resp ect to the earth's magnetic eld.
The instrumen t stays at the same altitude.
The instrumen t do es not receiv e an y mec hanical sho c k.
If the timebase ov en co ols (the instrumen t is disconnected from A C p ow er), y ou may ha v e to
readjust the timebase frequency after a new w armup cycle. T ypically , ho w ev er, the timebase
frequency returns to within 1 Hz of the original frequency .
The in ternal timebase can b e tested after reconnecting A C p ow er for 10
minutes, but for b est accuracy , test again after the instrumen t has b een on or
in standb y condition for 24 hours.
Agilent 8360 P erformance T ests
Description and Procedure

F requency c hanges due either to a c hange in orientation with resp ect to the earth's magnetic
eld, or to a c hange in altitude, usually go aw a y when the instrumen t is returned to its
original p osition. A frequency c hange due to mec hanical sho c k usually app ears as a xed
frequency error.
up for at least one hour.
If the oscilloscop e do es not ha v e a 50 input imp edance, connect c hannel 1
through a 50 feedthrough.

Digitizing Oscilloscop e HP/Agilent 54111D
F requency Standard HP/Agilent 5061A
2-8 P erformance T ests
Equipment

OUTPUT signal from the syn thesizer:
Channel 1:
Channel 2:
Timebase:
Delay Reference At center
Trigger:
Trig 3 Level 100 mV
Trigger Src Trig 3
Display:
Display Mode Real time
Agilent 8360 P erformance T ests

Aging Rate = (1 cycle/10 MHz) (1/T1 1/T3) (24 hours/T2)
T2 = 3 hours
T3 = 349 seconds
= (1 cycle/10 MHz) (1/351s 1/349s) (24h/3h)
= 1.306x10
If the absolute frequency of the standard and of the timebase oscillator are
extremely close, y ou can reduce the measuremen t time (T1 and T3) b y
measuring the time required for a phase c hange of less than 360
. In step 6,
c hange 1 cycle to 0.5 cycle for 180
, or 0.25 cycle for 90
In Case of Difficulty
conditions ha v e not c hanged throughout the test.
required for a 360
phase shift is to o short for an accurate measuremen t. If the 360
shift takes less than tw o minutes, p erform the \10 MHz Standard" adjustmen t.
Swept CW Gener ator T r oublesho oting Guide .
2-10 P erformance T ests

With the syn thesizer in swept mo de, the sp ectrum analyzer is set to zero span at the
measuremen t frequency . As the syn thesizer sweeps through the sp ectrum analyzer frequency
setting, a signal is generated on the sp ectrum analyzer's video output that is input to the
oscilloscop e.
The syn thesizer's TRIGGER OUTPUT, used to trigger the oscilloscop e, is a series of 1601
pulses, ev enly spaced during the sweep. The oscilloscop e is triggered on the pulse that
represen ts the desired measuremen t frequency , and the sp ectrum analyzer is tuned to display
the video output on the oscilloscop e.
up for at least one hour.

Figure 2-2. Sw ept Frequency Accuracy T est Setup
Sp ectrum Analyzer HP/Agilent 8566B
Digitizing Oscilloscop e HP/Agilent 54111D
Oscilloscop e Prob es HP/Agilent 10431A
T able 2-2. Then set:
Agilent 8360 P erformance T ests
Sw ept Frequency Accuracy
Description and Procedure · Equipment

T rigger 3 is a trigger enable that ensures that c hannel 2 (the true trigger)
triggers only on a forward sweep after the sp eci ed n um b er of ev en ts.
Channel 1:
Input Impedance 1 M
Channel 2:
Input Impedance 1 M
2-12 P erformance T ests
Sw ept Frequency Accuracy

Timebase:
Time/Division 25 s
Delay Reference At center
Trigger:
Trigger Src Chan 2
Trigger Level 1.6 V
Trigger Src Trig 3
Trigger Level 1.6 V
Trigger After Positive Edge
Trigger On Trig 3
Trigger On 33 events
Trigger Of Positive edge
Trigger On Channel 2
Display:
Display Mode Real time
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
USER CAL
Swp Span Cal Always (asterisk on)
First start frequency in T able 2-2
First stop frequency in T able 2-2
Set the frequency range rst, or y ou cannot set the sweep time prop erly .
SWEEP TIME
(F or Option 006, set to 150 ms)
F or the 83640B /L and 83650B /L only , also press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SERVICE
Disable Doubler (asterisk on)
Agilent 8360 P erformance T ests
Sw ept Frequency Accuracy

T able 2-2. Sw ept Frequency Accuracy Instrument Settings

oscilloscop e. (See Figure 2-3 .) Align the video input with the trigger output at cen ter
screen.
2-14 P erformance T ests
Sw ept Frequency Accuracy


Figure 2-3. Video Signal on the Oscilloscope
step 6 in T able 2-2 as frequency error.
through T able 2-6 as appropriate for eac h instrumen t mo del.
results in T able 2-2 and T able 2-3 through T able 2-6 :
x100 = p ercent
stop frequency start frequency
F or 83620B /22B/24B and 83623B /L only:
T able 2-3. Additional Instrument Settings
T rigger Ev ents
2 20 2.10125 10
2 20 6.8825 435
2 20 7.1525 459
2 20 13.25 1001
2 20 13.79 1049
2 20 19.775 1581
Agilent 8360 P erformance T ests
Sw ept Frequency Accuracy

F or 83630B /L only:
T able 2-4. Additional Instrument Settings

F or 83640B /L only:
T able 2-5. Additional Instrument Settings

2-16 P erformance T ests
Sw ept Frequency Accuracy

F or 83650B /L only:
T able 2-6. Additional Instrument Settings

YO Driver +10 V Reference
YO Driver Gain and Linearity
YO Delay Comp ensation
Sw eep Ramp Calibration
In Case of Difficulty
frequency with the syn thesizer's 10 MHz reference connected to the sp ectrum analyzer's
external reference.
Swept CW Gener ator T r oublesho oting Guide .
Agilent 8360 P erformance T ests
Sw ept Frequency Accuracy
Related Adjustments

The syn thesizer's Z-AXIS BLANK/MARKER output go es active high when a c hange in
frequency is initiated, and returns low when the syn thesizer settles at the new frequency .
Using an oscilloscop e to monitor this output, the frequency switching time is measured in
CW, step sweep, and frequency list mo des.
w arm up for at least one hour.

Figure 2-4. Frequency Switching Time T est Setup
Digitizing Oscilloscop e HP/Agilent 54111D
Oscilloscop e Prob es HP/Agilent 10431A
2-18 P erformance T ests
Description and Procedure · Equipment · Frequency Switching Time

CW Frequency Switching Time (Across Bandswitch P oints)
Channel 1:
Input Impedance 1 M
Channel 2:
Timebase:
Delay Reference At left
Delay 10 ms
Trigger:
Trigger Src Chan 1
Trigger Level 1.6 V
Display:
T able 2-7. CW Frequency Switching Time Settings
Initial CW F requency (GHz) Second CW F requency (GHz) Pulse Width
13.5 50
1 Set to 2 GHz (low est start frequency) for synthesizers not capable of 0.01 GHz.
2 Set to the highest stop frequency for synthesizers not capable of 50 GHz.
v alue as pulse width in T able 2-7 .
in T able 2-7.
Agilent 8360 P erformance T ests

Stepped Sw eep Frequency Switching Time (Within a Frequency Band)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
2.4
STOP
6.9
SWEEP
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MENU
SINGLE
Channel 1:
Timebase:
Delay Reference At left
Delay 2 ms
Trigger:
Trigger After Negative edge
Trigger Src Chan 1
Trigger On 1 events
Trigger Of Positive edge
Trigger On Channel 1
Display:
. The
oscilloscop e should display Awaiting Trigger.
.
Pulse width v alue .
Pulse width v alue .
Frequency List Frequency Switching Time (Across Bandswitch P oints)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNN
MENU
On the syn thesizer, press
Enter List Freq and en ter the following frequencies in the
frequency list men u:
2-20 P erformance T ests

F requency List for Syn thesizers with 0.01 GHz Capability
F requency (GHz):
Disregard if b ey ond y our syn thesizer's capabilit y .
F requency List for Syn thesizers without 0.01 GHz Capability
F requency (GHz):
Disregard if b ey ond y our syn thesizer's capabilit y .
Channel 1:
Channel 2:
Timebase:
Delay Reference At left
Delay 2 ms
Trigger:
Trigger After Negative edge
Trigger Src Channel 1
Trigger On 1 events
Trigger Of Positive edge
Trigger On Channel 1
Display:
Agilent 8360 P erformance T ests

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SWEEP
MENU
SINGLE
. The
oscilloscop e should display Awaiting Trigger.
.
Pulse Width V alue
1 ev en t
2 ev en ts
3 ev en ts
4 ev en ts
5 ev en ts
6 ev en ts
1 Not applicable for all mo dels. (Each even t corresp onds to a
change in frequency in the frequency list.)
and including 6.
F ractional-N V CO
YO Lo op Gain
In Case of Difficulty
adjust the trigger lev el on the oscilloscop e.
Swept CW Gener ator T r oublesho oting Guide .
2-22 P erformance T ests

Using a p ow er meter, c hec k the p ow er accuracy of the syn thesizer at sev eral CW frequencies.
A t eac h frequency , v erify that the actual output p ow er is within sp eci cation ov er the full
dynamic range of the ALC lo op.
on the syn thesizer.
a. Zero and calibrate the p ow er meter/sensor.
b. Set to dBm mo de.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
USER CAL
Peak RF Always (asterisk on).

Figure 2-5. P o w er Accuracy T est Setup
P ow er Meter HP/Agilent 436A/7A/8A
P ow er Sensor HP/Agilent 8487A (83640B /L and 83650B/L)
P ow er Sensor HP/Agilent 8485A (83620B /22B/24B, 83623B/L,
A tten uator 10 dB HP/Agilent 8490D Option 010 (83640B/L and 83650B/L)
A tten uator 10 dB HP/Agilent 8493C Option 010 (83620B/22B/24B,
En ter the rst CW v alue in T able 2-8 .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Uncoupl Atten (Option 001 only)
RF ON/OFF
O (amb er light o )
Agilent 8360 P erformance T ests
P o w er Accuracy

T able 2-8. P o w er Accuracy Frequencies
at 10 dBm
at > 10 dBm
1 Disregard if b eyond your synthesizer's capability .
measured.
syn thesizer. W rite this v alue do wn on a separate piece of pap er.
F or p ow er lev els of 10 dBm and ab ov e, measure the p ow er at 10 dBm, then set
the attenuator to the 10 dB p osition (use a 10 dB attenuator for syn thesizers
without a step attenuator.) Record the di erence b et w een the measuremen ts:
(approximately 10 dB). Then, with the attenuator still set to
10 dB, complete the remaining measuremen ts. Eac h of these measuremen ts
should b e o set b y the di erence in dB that y ou measured.
* k ey to incremen t the p ow er lev el 1 dB.
Record the w orst case measured di erence (greatest duration) for this frequency in
T able 2-8 for p ow er lev els 10 dBm and, if y our syn thesizer is capable, for p ow er lev els >
10 dBm.
maximum sp eci ed frequency of y our syn thesizer. Remem b er to delete the 10 dB of
attenuation for p ow er lev els b elow 0 dBm.
test record.
Mo dulator O set and Gain
ALC P ow er Lev el Accuracy
ALC P ow er Flatness
2-24 P erformance T ests
P o w er Accuracy

In Case of Difficulty
Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series
Swept CW Gener ator T r oublesho oting Guide .
Agilent 8360 P erformance T ests
P o w er Accuracy

This pro cedure uses the user atness correction arra y to automatically measure p ow er atness.
The p ow er meter is connected directly to the syn thesizer's RF output. The syn thesizer
con trols the p ow er meter via GPIB while the p ow er meter measures the RF output. (There
cannot b e another con troller on the GPIB during this test.) If the syn thesizer has a step
attenuator, it is set to 0 dB so that an y input in to the atness array indicates the RF output
p ow er atness.
This p erformance test requires an HP/Agilen t 437B p ow er meter. The correct
p ow er sensor calibration factors m ust b e loaded and selected.
This pro cedure deletes an y existing user atness correction arra y .
range b eing calibrated. Do not connect the p ow er sensor to the syn thesizer y et. Preset all
instrumen ts and let them w arm up for at least one hour.

Figure 2-6. P o w er Flatness T est Setup
P ow er Meter HP/Agilent 437A
P ow er Sensor HP/Agilent 8482A
P ow er Sensor HP/Agilent 8487A (83640B/L and 83650B/L)
P ow er Sensor HP/Agilent 8485A (83620B/22B/24B, 83623B/L,
Lo w Band P o w er Flatness
Disr e gar d if b eyond your synthesizer's c ap ability.
a. Delete all en tries in the user atness correction array . Press:
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
b. Set the auto ll stop frequency to 2 GHz:
Auto Fill Stop
2-26 P erformance T ests
P o w er Flatness

c. Set the auto ll start frequency to 15 MHz:
Auto Fill Start
d. Set the auto ll incremen t to 10 MHz:
Auto Fill Incr
a. Zero and calibrate the p ow er meter/sensor.
b. Connect the p ow er sensor to the syn thesizer's RF OUTPUT as sho wn in Figure 2-6 .
array for the frequency range en tered. Press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
When the atness correction array is completed, the men u is returned.
the maximum and minimum correction v alues. The di erence b et w een them is the p ow er
atness measuremen t. Record the v alue on the test record and compare the v alue to the
sp eci cation.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PO WER
MENU
High Band P o w er Flatness
A 20 GHz stop frequency .
A 2.05 GHz start frequency .
A 100 MHz incremen t.
Millimeter Band P o w er Flatness > 20 GHz and 40 GHz
Disr e gar d if b eyond your synthesizer's c ap ability.
A 40 GHz stop frequency (for 83630B /L only , set to 26.5 GHz).
A 20.05 GHz start frequency .
A 100 MHz incremen t.
Millimeter Band P o w er Flatness > 40 GHz
Disr e gar d if b eyond your synthesizer's c ap ability.
A 50 GHz stop frequency .
A 40.05 GHz start frequency .
A 100 MHz incremen t.
P ow er Flatness
In Case of Difficulty
Swept CW Gener ator T r oublesho oting Guide .
Agilent 8360 P erformance T ests
P o w er Flatness

This p erformance test is not v alid unless the syn thesizer meets b oth its p ow er
accuracy and p ow er atness sp eci cations. P erform those tests rst.
The unlev eled status indicator is display ed when the instrumen t is unlev eled as the syn thesizer
sweeps ov er sp eci c frequency ranges in fast con tinuous sweep, and fast and slow single sweep
op eration. Because of the syn thesizer's p ow er accuracy and atness p erformance, a p ow er
meter is not required for this measuremen t. The following pro cedure tests the most likely
w orst case situations for maximum lev eled p ow er.
w arm up for at least one hour.
The 20 dB attenuator provides a go o d match on the RF output. If the
syn thesizer has a step attenuator, y ou can simulate a go o d match b y
decoupling the step attenuator from the ALC and setting the attenuator to 20
dB.

Figure 2-7. Maximum Lev eled P o w er T est Setup
A tten uator 20 dB HP/Agilent 8490D Option 020 (83640B/L and 83650B /L)
A tten uator 20 dB HP/Agilent 8493C Option 020 (83620B/22B/24B,
83623B/L, and 83630B /L)
the RF OUTPUT with a go o d 50 ohm imp edance match suc h as a 20 dB attenuator or a
p ow er sensor (not necessary for syn thesizers with a step attenuator.) Press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MENU
W ait for the syn thesizer to complete auto tracking b efore con tinuing to the next step.
Set the p ow er to 1 dB b elow the sp eci ed maximum lev eled p ow er for the
syn thesizer's full frequency range.
On (amb er light is on)
CONT
SWEEP TIME
2-28 P erformance T ests
Maximum Lev eled P o w er

p ow er lev el un til the indicator just go es o . (Pow er is lev eled).
and initiate sev eral sweeps. If necessary , reduce the p ow er lev el
un til the unlev eled status indicator do es not turn on while sweeping.
0.5
Initiate sev eral sweeps. If necessary , reduce the p ow er lev el un til the unlev eled status
indicator do es not turn on while sweeping.
Record the p ow er lev el on the test record.
SWEEP TIME
listed. Repeat steps 3 through 7 for eac h of these new frequency ranges.
20.0 GHz (Option 001 only)
26.5 GHz
20.0 GHz (Option 001 only)
Ampli er/Multiplier Adjustmen ts
Ampli er/Filter Adjustmen ts
Mo dulator O set and Gain
In Case of Difficulty
Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series
Swept CW Gener ator T r oublesho oting Guide .
Agilent 8360 P erformance T ests
Maximum Lev eled P o w er

The syn thesizer is set up to externally lev el using a negative crystal detector. A DVM
measures the crystal detector DC output. A 1000 pf capacitor across the DVM input limits
the e ect of an y small signals generated b y the DVM. The syn thesizer p ow er lev el is set to the
minimum allo w able limit and then is increased un til the p ow er output is lev eled. The external
lev eling v oltage is then measured and compared to the minimum sp eci cation. Next, the
output p ow er is increased un til the maxim um lev eled p ow er is reac hed. The lev eling v oltage is
measured and compared to the sp eci cation.
The v oltage , not the p ow er lev el , is what is imp ortant in p erforming this test.
If y ou ha v e problems ac hieving the sp eci ed v oltages in this test, y ou may place a pad in the
RF path b efore the crystal detector or the coupling factors may b e c hanged to ac hiev e the
sp eci ed v oltages.
The syn thesizer m ust meet the p ow er atness sp eci cation b efore pro ceeding
with this test.

Figure 2-8. External Lev eling T est Setup
Digital V oltmeter HP/Agilent 3456A
Crystal Detector HP/Agilent 33330E (83640B /L and 83650B/L)
Crystal Detector HP/Agilent 33330D (83620B/22B/24B,
Capacitor 1000 pf P/N 0160-4574
RF output with a go o d 50 imp edance match suc h as a 10 dB attenuator or a p ow er
sensor (not necessary for syn thesizers with a step attenuator). Press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
USER CAL
Auto Track (asterisk on)
W ait for the syn thesizer to complete auto tracking b efore con tinuing to the next step.
2-30 P erformance T ests

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ALC
POWER LEVEL
resolution and the n um b er of in tegration cycles. On the HP/Agilent 3456A, set:
N CYC INT
STORE
N DIG DISP
Set the DVM to measure DC v olts.
already out, leav e the p ow er lev el at 36 dBm.)
Record the DVM reading on the test record and compare the reading to the sp eci cation.
If the UNLVLED message turns on b efore reac hing +4 dBm, set the stop
frequency to a low enough frequency where the sp eci ed output p ow er
is 4 dBm. If y ou don't get an unlev eled message, disconnect the BNC
tee temp orarily to make sure that the unlev eled message turns on. Then
reconnect the equipmen t and con tinue with the test.
. Increase the syn thesizer p ow er lev el un til the UNLVLED message turns
on. Decrease p ow er just un til the message go es out.
Ampli er/Multiplier Adjustmen ts
Ampli er/Filter Adjustmen ts
P ow er Flatness
In Case of Difficulty
calibration (under the
k ey .)
Swept CW Gener ator T r oublesho oting Guide .
Agilent 8360 P erformance T ests

Use this pro cedure to measure the syn thesizer's harmonics and subharmonics ov er its en tire
frequency range. Harmonics are in teger m ultiples of the syn thesizer RF output frequency and
subharmonics are fractional m ultiples of the YO frequency . Subharmonics do not exist at all
frequencies. In low band, mixer spurs and fundamen tal feedthrough are the most signi cant
harmonically-related signals. The mixer spur/YO frequency relationship is as follows:
X (5.4 GHz xed oscillator) Y (YO frequency) = Mixer Spur F requency
where X and Y are in tegers.
Other harmonics are typically direct m ultiples of the YO frequency .
In this pro cedure, the syn thesizer is man ually swept ov er its frequency range while the
sp ectrum analyzer measures the harmonics and subharmonics in eac h frequency band. An y
harmonics or subharmonics that are within 5 dB of the sp eci cation are subsequen tly v eri ed
with a more accurate pro cedure.
let them w arm up for at least one hour.
On the HP/Agient 8566B, connect the CAL OUTPUT to the RF INPUT. Press
W
to calibrate the sp ectrum analyzer.

Figure 2-9. Spurious Signals (Harmonic) T est Setup < 20 GHz
Sp ectrum Analyzer HP/Agilent 8566B
A tten uator 10 dB HP/Agilent 8490D Option 010 (83640B/L
A tten uator 10 dB HP/Agilent 8493C Option 010 (83620B /22B/24B,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Peak RF Always (asterisk on).
2-32 P erformance T ests
Spurious Signals (Harmonics Subharmonics)

Harmonic Measurement < 20 GHz
Set the maximum sp eci ed lev eled p ow er.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SWEEP
MENU
Then set:
Video Bandwidth 3 MHz
T able 2-9. Start and Stop Frequencies
2.0 7.0
7.0 13.5
13.5 20.0
1 Set to 2 GHz for 83622B/24B.
analyzer display for harmonics and subharmonics.
See T able 2-10 for the YO frequency ranges that corresp ond to the RF output frequencies.
T able 2-10.
Corresponding YO Frequency Ranges and RF Output Frequencies
5.41 to 7.4 N/A 0.01 to 2.0
2.0 to 7.0 1 2.0 to 7.0
3.5 to 6.75 2 7.0 to 13.5
4.5 to 6.67 3 13.5 to 20.0
an y harmonic is within 5 dB of sp eci cation, make a more accurate measuremen t using
the \Harmonic/Subharmonic V eri cation Pro cedure < 20 GHz" that follo ws.
case harmonic for RF output frequencies of 0.01 to 2.0 GHz.
Agilent 8360 P erformance T ests
Spurious Signals (Harmonics Subharmonics)

F or the 83624B , 83623B /L, 83640B /L, and 83650B /L and all for instrumen ts with Option
frequencies of 10 to < 50 MHz and 50 MHz to 2.0 GHz.
2-34 P erformance T ests
Spurious Signals (Harmonics Subharmonics)

syn thesizer across the frequency range and c hec k the sp ectrum analyzer display for
harmonics and subharmonics.
and v erify an y that are within 5 dB of sp eci cation with the \Harmonic/Subharmonic
V eri cation Pro cedure < 20 GHz."
harmonic for RF output frequencies of 2.0 and 20 GHz and the v alue of the w orst
case subharmonic for RF output frequencies of 7 to 20 GHz.
F or the 83630B /L, 83640B /L, and 83650B /L: Record, on the test record, the w orst case
subharmonic for RF output frequencies of 7 and 20 GHz. Note, for future use, the
w orst case harmonic for RF output frequencies of 2.0 to 20 GHz .
Harmonic Measurement 20 to 26.5 GHz
Disr e gar d if b eyond your synthesizer's c ap ability.
to 26.5 GHz).

Figure 2-10. Spurious Signals (Harmonics) T est Setup > 20 GHz
Sp ectrum Analyzer HP/Agilent 8566B
Micro w av e Ampli er HP/Agilent 11975A
Harmonic Mixer (20 to 26.5 GHz) HP/Agilent 11970K
Harmonic Mixer (26.5 to 40 GHz) HP/Agilent 11970A
Harmonic Mixer (40 to 50 GHz) HP/Agilent 11970Q
1 Firm w are revision 13.8.86 or later.
Agilent 8360 P erformance T ests
Spurious Signals (Harmonics Subharmonics)
Equipment

range and set the reference lev el o set to comp ensate for the mixer's con v ersion loss at 23
GHz.
Adjust the reference lev el to p osition the signal p eak at the top reference graticule. Use
this reference lev el for all harmonic measuremen ts.
analyzer display for harmonics and subharmonics. See T able 2-11 for the YO frequency
ranges that corresp ond to the RF output frequencies.
Since an external mixer is used, use the sp ectrum analyzer signal iden tify feature to v erify
that susp ected signals are in the 20 to 26.5 GHz frequency range. The signals most likely
to app ear are the fth and sixth YO harmonics.
the test record. If an y harmonics or subharmonics are within 5 dB of the sp eci cation,
make a more accurate measuremen t using the \Harmonic/Subharmonic V eri cation
Pro cedure > 20 GHz" that follo ws. Note the w orst case harmonic and the
w orst case subharmonic from 20 to 26.5 GHz.
to 26.5 GHz.
F or the 83630B /L, 83640B /L, and 83650B /L only: F rom steps 12 and 20, determine the
w orst case harmonic from 2.0 to 26.5 GHz and record this v alue on the test record.
T able 2-11.
Corresponding YO Frequency Ranges and RF Output Frequencies
Mo del YO F requencies
5.0 to 6.625 4 4 20 to 26.5
83650B /L 4.41667 to 5.5667 3 6 26.5 to 33.4
83650B /L 5.5667 to 6.333 3 6 33.4 to 38
83650B /L 4.75 to 6.25 4 8 38 to 50
83640B /L 5.0 to 6.375 2 4 20 to 25.5
83640B /L 4.25 to 5.0 3 6 25.5 to 30
83640B /L 5.0 to 6.667 3 6 30 to 40
2-36 P erformance T ests
Spurious Signals (Harmonics Subharmonics)

Harmonic Measurement 26.5 to 40 GHz
Disr e gar d if b eyond your synthesizer's c ap ability.
(26.5 to 40 GHz).
range and set the reference lev el o set to comp ensate for the mixer's con v ersion loss at 33
GHz.
Adjust the reference lev el to p osition the signal p eak at the top reference graticule. Use
this reference lev el for all harmonic measuremen ts.
analyzer display for harmonics and subharmonics. See T able 2-11 for the YO frequency
ranges that corresp ond to the RF output frequencies.
Since an external mixer is used, use the sp ectrum analyzer signal iden tify feature to v erify
that susp ected signals are in the 26.5 to 40 GHz frequency range. The signals most likely
to app ear are the fth and sixth YO harmonics.
the test record. If an y harmonics or subharmonics are within 5 dB of the sp eci cation,
make a more accurate measuremen t using the \Harmonic/Subharmonic V eri cation
Pro cedure > 20 GHz" that follo ws. Note the w orst case harmonic and the
w orst case subharmonic from 26.5 to 40 GHz.
40 GHz and record this v alue on the test record.
F or the 83640B /L/ and 83650B /L only: F rom steps 20 and 28, determine the w orst case
subharmonic from > 20 and 40 GHz and record this v alue on the test record.
Harmonic Measurement 38 to 50 GHz
Disr e gar d if b eyond your synthesizer's c ap ability.
GHz).
set the reference lev el o set to comp ensate for the con v ersion loss at 40 GHz.
Agilent 8360 P erformance T ests
Spurious Signals (Harmonics Subharmonics)

Adjust the reference lev el to p osition the signal p eak at the top reference graticule. Use
this reference lev el for all harmonic measuremen ts.
analyzer display for harmonics and subharmonics. See T able 2-11 for the YO frequency
ranges that corresp ond to the RF output frequencies.
Since an external mixer is used, use the sp ectrum analyzer signal iden tify feature to v erify
that susp ected signals are in the 38 to 50 GHz frequency range. The signals most likely to
app ear are the fth and sixth YO harmonics.
the test record. If an y harmonics or subharmonics are within 5 dB of the sp eci cation,
make a more accurate measuremen t using the \Harmonic/Subharmonic V eri cation
Pro cedure > 20 GHz" that follo ws. Note the w orst case harmonic and the
w orst case subharmonic from 38 to 50 GHz.
26.5 to 50 GHz and record this v alue on the test record. F rom step 36, determine the
w orst case subharmonic from > 40 and 50 GHz and record this v alue on the test record.
Harmonic/Subharmonic V erification Procedure < 20 GHz
subharmonic to b e measured.
2-38 P erformance T ests
Spurious Signals (Harmonics Subharmonics)

Measure the signal lev el with the sp ectrum analyzer marker. Change the reference lev el as
necessary .
the carrier, as follo ws:
Harmonic/Subharmonic amplitude (dBc) = [Syn thesizer RF output amplitude (dBm)
F or example:
RF Output = +10 dBm
Harmonic = 60 dBm
Harmonic (dBc) = [RF Out (Harmonic)]
= [10 (60 dBm)]
= [10 + 60 dBm]
= 70 dBc
Harmonic/Subharmonic V erification Procedure > 20 GHz
F or accurate measuremen ts, the syn thesizer m ust meet its p ow er atness
sp eci cation.
measured.
Agilent 8360 P erformance T ests
Spurious Signals (Harmonics Subharmonics)
(Harmonic/Subharmonic amplitude (dBm))]

necessary .
the carrier, as follo ws:
Harmonic/Subharmonic Amplitude (dBc) = [Syn thesizer RF output amplitude (dBm)
F or example:
RF Output = +10 dBm
Harmonic = 60 dBm
Harmonic (dBc) = [RF Out (Harmonic)]
= [10 (60 dBm)]
= [10 + 60 dBm]
= 70 dBc
In Case of Difficulty
Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series
Swept CW Gener ator T r oublesho oting Guide .
2-40 P erformance T ests
Spurious Signals (Harmonics Subharmonics)
Harmonic/Subharmonic amplitude (dBm)] · Related Adjustments · None

Use this pro cedure to measure kno wn, xed, o set spurs that are generated in the frequency
syn thesis section of the syn thesizer. The syn thesizer is set to v arious CW frequencies where
these spurious signals will most likely o ccur. Then the sp ectrum analyzer is tuned to the spur
frequencies to measure their lev els.
On the HP/Agilent 8566B, connect the CAL OUTPUT to the RF INPUT. Press
W
to calibrate the sp ectrum analyzer.

Figure 2-11. Spurious Signals (Non-Harmonic) T est Setup
Sp ectrum Analyzer HP/Agilent 8566B
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
USER CAL
Peak RF Always (asterisk on)
19.765
Set the maximum sp eci ed lev eled p ow er.
Agilent 8360 P erformance T ests

to cen ter frequency . Decrease the frequency span to 100 Hz k eeping the signal cen tered
on the display and then rep eat the marker p eak searc h and set the marker to cen ter
frequency .
whic h the spurs are referenced.
a.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Select Cal .
b. Use the RPG knob to select the rst calibration constant in T able 2-12 (#18).
c. T o the v alue of the sp ectrum analyzer cen ter frequency (19.765 GHz), add the v alue of
the calibration constant (or twice the v alue, as indicated in T able 2-12) and the v alue
from the second column of T able 2-12 (20 kHz). Record this v alue (in Hz) in the last
column of T able 2-12 .
For example:
19.765 GHz (Spectrum analyzer center frequency)
+ 338 kHz (Possible value of calibration constant #18
+ 20 kHz (Value from second column of the table)
= 19,765,358,000 Hz (Record this result in the last column
of the table.)
Repeat this pro cedure for the remaining calibration constants in T able 2-12. The
last tw o actual spur frequencies ha v e already b een calculated and are not related to
calibration constants.
T able 2-12. Determining the Actual Spur Frequencies
Calibration Constan t #18 20
2(Calibration Constan t #18) 40
Calibration Constan t #19 60
2(Calibration Constan t #19) 120
N/A 500 19,765,500,000
from T able 2-12 . Then set:
2-42 P erformance T ests

use the marker to measure its amplitude. If the spur is in the noise lev el, use the noise
lev el amplitude (this gives a w orst case v alue).
dBm Spur Absolute Amplitude
Carrier Amplitude (dBm) Spur Absolute Amplitude (dBm) = Spur Lev el (dBc)
Record the result on the test record. Compare the result to the sp eci cation.
T able 2-12.
Lo w Band Offset Spurious Signals
Disr e gar d if b eyond your synthesizer's c ap ability.
POWER LEVEL
carrier amplitude to whic h the spurs are referenced.
dBm RF output at 1 GHz
the marker to measure the spurs (c hange the reference lev el as necessary).
Calculate the spur lev el in dBc for eac h of the spurs as follows:
Spur Lev el (dBc) = [RF Output at 1 GHz (dBm) (Spur Lev el)]
Record the spur lev el in dBc on the test record. Compare the spur lev el to the
sp eci cation.
T able 2-13. Lo w Band Spurious Signals
Sp ectrum Analyzer F requency Spur
100 MHz 100 MHz Fixed Spur
900 MHz 100 MHz O set Spur
6.4 GHz LO F eedthrough Spur
Agilent 8360 P erformance T ests

Lo w Band Mixer Spurs
1.6
results in the most dominant mixing spur moving from 1.4 to 1.8 GHz. The spur should
b e at the left-hand edge of the display (1.4 GHz). It should then trav el to the right as
y ou tune do wn. If y ou can't see it, increase p ow er un til y ou can iden tify the spur. Then
reset the p ow er lev el to 0 dBm (the p erformance sp eci cation is v alid for 0 dBm). Y ou are
nished tuning when the spur merges with the 2 GHz carrier.
in the noise o or use that v alue, and record the di erence b et w een the reference lev el and
spurious signal amplitude on the test record.
F ractional-N Reference and API Spurs
In Case of Difficulty
Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series
Swept CW Gener ator T r oublesho oting Guide .
2-44 P erformance T ests

Use this pro cedure to measure line-related spurs. The syn thesizer is placed at a CW frequency
where the syn thesizer is most sensitive to line-related spurs. The sp ectrum analyzer measures
the RF output o set b y the harmonics of the line p ow er frequency (60 Hz or 50 Hz). T o
eliminate measuring the sp ectrum analyzer line-related spurs, an In v ertron is used to op erate
the sp ectrum analyzer at a di eren t line frequency so that its own spurs will not a ect the
measuremen t.
The sp ectrum analyzer m ust ha v e the capabilit y of a 10 Hz resolution bandwidth. A wider
resolution bandwidth results in the syn thesizer's phase noise masking the spur.
w arm up for at least one hour.

Figure 2-12. Spurious signals (Line-Related) T est Setup
Sp ectrum Analyzer HP/Agilent 8566B
In v ertron California Instrumen ts 501TC
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Peak RF Always (asterisk on)
6.99
Agilent 8360 P erformance T ests

di erence of the marker from the cen ter frequency . Calibrate the sp ectrum analyzer
frequency o set so that the signal p eak is in the cen ter of the display .
whic h the spurs are referenced.
This pro cedure is written for a line frequency of 60 Hz. F or a 50 Hz line
frequency , lo ok for spurs at frequencies of 50 Hz times the harmonics in
T able 2-14, and set the sp ectrum analyzer frequencies accordingly .
T able 2-14. Then set:
use the marker to measure its amplitude. If the spur is in the noise lev el, use the noise
lev el amplitude (this gives a w orst case v alue).
dBm Spur Absolute Amplitude
Spur Lev el (dBc) = [Carrier Amplitude (dBm) (Spur Lev el)]
Record the result on the test record. Compare the result to the sp eci cation.
The o dd harmonics will normally b e higher than the ev en harmonics.
T able 2-14. Change the sp ectrum analyzer reference lev el as indicated in the table.
T able 2-14.
Spectrum Analyzer Frequencies and Line Spur Frequencies and Harmonics
2 120 6,990,000,120 50
3 180 6,990,000,180 50
4 240 6,990,000,240 50
analyzer to 1.9 GHz and rep eat steps 6 through 11 for this frequency .
2-46 P erformance T ests

In Case of Difficulty
Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series
Swept CW Gener ator T r oublesho oting Guide .
Agilent 8360 P erformance T ests
Related Adjustments · None · Spurious Signals (Line-Related)

This pro cedure uses a phase noise measuremen t system. It is the fastest and most accurate
metho d for measuring phase noise.
up for at least one hour.
2-48 P erformance T ests
Single Sideband Phase Noise
Description and Procedure


Figure 2-13. Single Sideband Phase Noise T est Setup
Phase Noise Measuremen t System HP/Agilent 3048A
Agilent 8360 P erformance T ests
Single Sideband Phase Noise
Equipment

6.99
T o av oid en tering these parameters eac h time y ou p erform this test, create a
le on the phase noise measuremen t system con taining these parameters.
With these parameters en tered, the softw are automatically sets the remaining parameters.
T able 2-15 sho ws the complete set of parameters.
T able 2-15. Phase Noise Measurement System P arameters
in time domain (like an oscilloscop e). The other trace is in the frequency domain (like a
sp ectrum analyzer).
2-50 P erformance T ests
Single Sideband Phase Noise

A t the message VERIFY BEATNOTE
w av e frequency with the frequency domain signal near the left edge (0 Hz).
to the sp eci cation.
T able 2-16. Frequency Setting for Phase Noise Measurements
Phase Noise Measurement System
1.97 310
18.0 18.0 720
1 Disregard if b eyond your synthesizer's capabilit y .
In Case of Difficulty
should b e ignored.
zero b eat can b e found.
and refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series
Swept CW Gener ator T r oublesho oting Guide .
Agilent 8360 P erformance T ests
Single Sideband Phase Noise

(F or B-Series Only)
Using a sp ectrum analyzer, the syn thesizer's CW RF output p ow er is measured b oth with
pulse on and with pulse o . The di erence in p ow er is the pulse on/o ratio.
w arm up for at least one hour.
On the HP/Agilent 8556B, connect the CAL OUTPUT to the RF INPUT. Press
to calibrate the sp ectrum analyzer.

Figure 2-14. Pulse Modulation On/Off Ratio T est Setup
Sp ectrum Analyzer HP/Agilent 8566B
2-52 P erformance T ests
Pulse Modulation On/Off Ratio

The rst cen ter frequency in T able 2-17.
F or all other syn thesizers: Set the p ow er lev el to 9.9 dBm b y pressing:
9.9
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Up/Dn Size CW
T able 2-17. Pulse On/Off Center Frequencies
Cen ter F requency (GHz) Pulse On/O Ratio
1 Disregard if b ey ond your synthesizer's capability .
a. T une cen ter frequency to cen ter the signal on the display .
b. Set the marker to highest p eak.
c. Note the marker amplitude.
Pulse On/Off Extrnl (asterisk on)
or, for syn thesizers with Option 002, press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Pulse On/Off Ext (asterisk on)
pulse on/o ratio. Record this v alue in T able 2-17.
MOD
Pulse On/Off Extrnl (asterisk o )
or, for syn thesizers with Option 002, press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Pulse On/Off Ext (asterisk o )
frequencies in T able 2-17 with the following c hanges to step 7 when testing 6.0 GHz:
Agilent 8360 P erformance T ests
Pulse Modulation On/Off Ratio

F or a CW F requency of 6.0 GHz:
a. Step the syn thesizer and sp ectrum analyzer in 50 MHz steps from 6.0 to 7.0 GHz.
b. Note the frequency of the highest amplitude signal. Set the syn thesizer and sp ectrum
analyzer to this frequency .
c. Set the marker to the highest p eak and note the maximum marker amplitude.
In Case of Difficulty
analyzer and the syn thesizer b y 1 kHz, and retest at that p oint. If the on/o ratio is bad,
it will b e bad ov er greater than a 100 kHz region.
Swept CW Gener ator T r oublesho oting Guide .
2-54 P erformance T ests
Pulse Modulation On/Off Ratio
Related Adjustments · None

(F or B-Series Only)
In this pro cedure, the syn thesizer's RF output is display ed directly on a high-bandwidth
digitizing oscilloscop e where the rise and fall times are measured and a comparison of pulsed
and CW p ow er lev el accuracy is made. P ow er lev el accuracy is c hec k ed at a 10 Hz pulse
rep etition rate to v erify its op eration at the slow est sp eci ed rate.
Rise and F all Times
w arm up for at least one hour.

Figure 2-15. Pulse P erformance T est Setup
Digitizing Oscilloscop e HP/Agilent 54124T
Pulse Generator HP/Agilent 8112A
A tten uator 10 dB HP/Agilent 8490D Option 010 (83640B/50B)
A tten uator 10 dB HP/Agilent 8493C Option 010 (83620B/22B/23B/24B/30B)
A tten uator 20 dB HP/Agilent 8490D Option 020 (83640B/50B)
A tten uator 20 dB HP/Agilent 8493C Option 020 (83620B/22B/23B/24B/30B)
First syn thesizer frequency in T able 2-18.
POWER LEVEL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ALC
Leveling Point Internal (asterisk on)
Agilent 8360 P erformance T ests

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Leveling Mode Search (asterisk on)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
USER CAL
Peak RF Always (asterisk on)
T able 2-18. Synthesizer Frequencies
Risetime F alltime Lev el
1 Disregard if b eyond your synthesizer's capability .
The CW frequencies c hec k ed provide a minimum v eri cation of pulse
p erformance. If pulse p erformance at a di eren t frequency is imp ortant,
rep eat the test at that CW frequency .
Channel 3:
Volts/Division 15 mV (Readjust as required)
Timebase:
Delay Reference At center
Trigger:
Trigger Level 1.2 V
Display:
2-56 P erformance T ests

Display Time 1 sec
On the oscilloscop e, turn the w av eform math function on. Then use the
w av eform math maxim um function to determine the pulse en v elop e when
making the follo wing measuremen ts.
MOD
Pulse On/Off Extrnl (asterisk on)
or, for syn thesizers with Option 002, press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Pulse On/Off Ext (asterisk on)
a. Adjust the timebase dela y to p osition the rising edge of the pulsed RF near the cen ter
of the display .
b. Adjust the c hannel 3 v olts/division and o set to obtain a 5 division signal lev el
b et w een the RF p ow er o and the RF p ow er on. (See Figure 2-16 .)
c. Adjust the c hannel 3 o set to mov e the RF p ow er o line one-half division b elow the
nearest horizontal graticule (making the horizontal graticule cross the w av eform at the
10% p oint.)
d. Adjust the oscilloscop e timebase to widen the pulse en v elop e so that the measuremen t
can b e made with greater accuracy . (See Figure 2-17 .)
e. Use the oscilloscop e's delta T function to measure the time di erence b et w een the 10%
and 90% risetime p oints on the en v elop e. (The 10% p oint of the risetime is where the
pulse en v elop e crosses the graticule. The 90% p oint is 4 divisions up.) Record this
v alue in T able 2-18.

Figure 2-16. Pulse Env elope Display ed Ov er 5 Divisions
Agilent 8360 P erformance T ests

oscilloscop e display .
v alue in T able 2-18 .

Figure 2-17. Widened Pulse Env elope
Pulse Lev eling Accuracy
First syn thesizer frequency in T able 2-18
POWER LEVEL
ALC
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MOD
Pulse On/Off Extrnl (asterisk on)
or, for syn thesizers with Option 002, press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Pulse On/Off Ext (asterisk on)
2-58 P erformance T ests

Channel 3:
Timebase:
Delay Reference At left
Trigger:
Trigger Level 50 mV
Increase the oscilloscop e sensitivit y to 5 mV/division. Adjust the dela y and timebase so
that the pulse w av eform takes up the en tire display . Adjust the oscilloscop e c hannel 3
o set to p osition the top of the pulse en v elop e near the cen ter graticule.
Pulse On/Off Ext (asterisk o )
.3
lev el (if the w av eform math function w as previously on, clear the display b efore executing
this step.) Use the oscilloscop e's delta V feature and p osition marker 1 at the cen ter of
the p eak p ow er as display ed b y the math w av eform (the width of the w av eform is caused
b y noise in the system.) Lea v e the marker at that p osition.
to determine the p eak p ow er lev el. Use the oscilloscop e's delta V feature and p osition
marker 2 at the cen ter of the p eak p ow er as display ed b y the math w av eform.
If the di erence b et w een marker 1 and marker 2 is less than 1 division, rep eat steps 13
through 17 with c hannel 3 v olts/division set to a more sensitive scale.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Pulse On/Off Extrnl (asterisk on)
16 and 18. Record the \Pass" or \F ail" in the Lev el Accuracy column in T able 2-18.
F ail , record \F ail" on the test record.
In Case of Difficulty
Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series
Swept CW Gener ator T r oublesho oting Guide .
Agilent 8360 P erformance T ests

(F or B-Series Only)
In this pro cedure, the syn thesizer's RF output is do wncon v erted in frequency so that an
oscilloscop e can measure the rise and fall times and make a comparison of pulsed and CW
p ow er lev el accuracy . F or measuremen ts ab ov e 20 GHz RF output frequency , a sp ectrum
analyzer with external mixer is used to do wncon v ert to a 321.4 MHz RF. P ow er lev el accuracy
is c hec k ed at a 10 Hz pulse rep etition rate to v erify its op eration at the slow est sp eci ed
rate. The p ow er ampli er ensures a sucien t signal lev el in to the oscilloscop e to make the
measuremen t.
let them w arm up for at least one hour.
Rise and F all Times < 20 GHz
This pro cedure is not v alid for fast pulse option (Option 006).
2-60 P erformance T ests
Pulse P erformance (Alternate)
Description and Procedure


Figure 2-18. Pulse P erformance T est Setup < 20 GHz
(Syn thesized Sw eeper) 83620A/B
Digitizing Oscilloscop e HP/Agilent 54111D
Preampli er HP/Agilent 8447F
A tten uator 6 dB HP/Agilent 8493C Option 006
A tten uator 10 dB HP/Agilent 8490D Option 010
A tten uator 10 dB HP/Agilent 8493C Option 010
Agilent 8360 P erformance T ests
Pulse P erformance (Alternate)
Equipment · Lo cal Oscillator · Mixer P/N 0955-0307

First syn thesizer frequency in T able 2-19 .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Leveling Point Internal (asterisk on)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Leveling Mode Search (asterisk on)
T able 2-19. Then set:
T able 2-19. Synthesizer and LO Frequencies
Risetime F alltime Lev el
5.0 4.5
9.0 8.5
15.0 14.5
1 Disregard if b eyond your synthesizer's capabilit y .
The CW frequencies c hec k ed provide a minimum v eri cation of pulse
p erformance. If pulse p erformance at a di eren t frequency is imp ortant,
rep eat the test at that CW frequency .
2-62 P erformance T ests
Pulse P erformance (Alternate)

Channel 1:
Input Impedance 50 ohms
Channel 2:
Timebase:
Delay Reference At center
Trigger:
Trigger Source Chan 2
Trigger Level 2 V
Display:
Number of Averages 1
On the oscilloscop e, turn the w av eform math function on. Then use the
w av eform math maxim um function to determine the pulse en v elop e when
making the follo wing measuremen ts.
Pulse On/Off Extrnl (asterisk on)
Agilent 8360 P erformance T ests
Pulse P erformance (Alternate)

a. Adjust the timebase dela y to p osition the rising edge of the pulsed RF near the cen ter
of the display .
b. Adjust the c hannel 1 v olts/division and o set to obtain a 5 division signal lev el
b et w een the RF p ow er o and the RF p ow er on (see Figure 2-19).
c. Adjust the c hannel 1 o set to mov e the RF p ow er o line one-half division b elow the
nearest horizontal graticule (making the horizontal graticule cross the w av eform at the
10% p oint.)
d. Use the oscilloscop e's delta T function to measure the time di erence b et w een the 10%
and 90% risetime p oints on the en v elop e. (The 10% p oint of the risetime is where the
pulse en v elop e crosses the graticule. The 90% p oint is 4 divisions up.) Record this
v alue in T able 2-19.

Figure 2-19. Signal Lev el betw een RF P o w er Off and On
oscilloscop e display .
this v alue in T able 2-19.
v alue from T able 2-19 on the appropriate test record.
Pulse Lev eling Accuracy < 20 GHz
2-64 P erformance T ests
Pulse P erformance (Alternate)

First syn thesizer frequency in T able 2-19
POWER LEVEL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ALC
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Pulse On/Off Extrnl (asterisk on)
Channel 1:
Timebase:
Delay Reference At left
Adjust the dela y and timebase so that the pulse w av eform takes up the en tire display .
Adjust the oscilloscop e c hannel 1 o set to p osition the top of the pulse en v elop e near the
cen ter graticule.
Pulse On/Off Extrnl (asterisk o )
0.3
Agilent 8360 P erformance T ests
Pulse P erformance (Alternate)

lev el (if the w av eform math function w as previously on, clear the display b efore executing
this step.) Use the oscilloscop e's delta V feature and p osition marker 1 at the cen ter of
the p eak p ow er as display ed b y the math w av eform (the width of the w av eform is caused
b y noise in the system.) Lea v e the marker at that p osition.
0.3
to determine the p eak p ow er lev el. Use the oscilloscop e's delta V feature and p osition
marker 2 at the cen ter of the p eak p ow er as display ed b y the math w av eform.
If the di erence b et w een marker 1 and marker 2 is less than 1 division, c hange c hannel 1
v olts/division to a more sensitive scale and rep eat steps 16 through 20.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MOD
Pulse On/Off Extrnl (asterisk on)
p ow er. It should stay within the tw o delta V markers set in steps 17 and 19. Record the
v alue in T able 2-19 .
T able 2-19 on the appropriate test record.
2-66 P erformance T ests
Pulse P erformance (Alternate)

Rise and F all Times > 20 GHz
Disr e gar d if b eyond your synthesizer's c ap ability.

Figure 2-20. Pulse P erformance T est Setup >20 GHz
Sp ectrum Analyzer HP/Agilent 8566B
Digitizing Oscilloscop e HP/Agilent 54111D
Micro w av e Ampli er HP/Agilent 11975A
RF Ampli er HP/Agilent 8447F
Pulse Generator HP/Agilent 8116A
Harmonic Mixer HP/Agilent 11970K
Harmonic Mixer HP/Agilent 11970A (83640B/50B)
Harmonic Mixer HP/Agilent 11970Q (83650B)
A tten uator 6 dB HP/Agilent 8493C Option 006
A tten uator 10 dB HP/Agilent 8490D Option 010
Agilent 8360 P erformance T ests
Pulse P erformance (Alternate)
Equipment

T able 2-20. Risetime and F alltime V alues
CW F requency Risetime F alltime Lev el Accuracy
Procedure > 20 to 26.5 GHz.
POWER LEVEL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Leveling Point Intrnl (asterisk on)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Leveling Mode Search (asterisk on)
mixer. Then set:
83630B only , record the w orst case risetime and w orst case falltime v alues from T able 2-19
and T able 2-20 on the test record.
Procedure > 26.5 to 40 GHz.
35 GHz, recording the risetime and falltime v alues in T able 2-20. F or 83640B only , record
the w orst case risetime and w orst case falltime v alues from T able 2-19 and T able 2-20 on
the test record.
Procedure > 40 GHz to 50 GHz.
GHz).
45 GHz and record the risetime and falltime v alues in T able 2-20. F or the 83650B
only , record the w orst case risetime and w orst case falltime v alues from T able 2-19 and
T able 2-20 on the test record.
Pulse Lev eling Accuracy > 20 GHz
Disr e gar d if b eyond your synthesizer's c ap ability.
2-68 P erformance T ests
Pulse P erformance (Alternate)

Procedure > 20 to 26.5 GHz.
the test record.
Procedure > 26.5 to 40 GHz.
accuracy in T able 2-20.
the test record.
Procedure > 40 to 50 GHz.
in T able 2-20.
the test record.
Agilent 8360 P erformance T ests
Pulse P erformance (Alternate)

In Case of Difficulty
to v erify a problem with the ampli er.
compression). The 0.3 dB v ariation should b e approximately equal on b oth sides. T ry a
low er syn thesizer p ow er lev el to get the ampli er out of saturation.
Swept CW Gener ator T r oublesho oting Guide .
2-70 P erformance T ests
Pulse P erformance (Alternate)
Related Adjustments · None

(F or B-Series Only)
In CW, at sp eci ed maximum lev eled p ow er, the syn thesizer is pulse mo dulated using a
pulse generator. The syn thesizer's RF output is ltered (only the video feedthrough passes),
ampli ed, and display ed on an oscilloscop e. Because of the low amplitude of the video
feedthrough, a preampli er is used in the test setup; system gain m ust b e considered when
making this measuremen t.
When video feedthrough is expressed as a p ercen tage, the follo wing equation is used:
Video feedthrough (%) = [(video Vp/carrier Vp) x 100]/video gain
hour.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Peak RF Always (asterisk on)
directly to c hannel 1 of the oscilloscop e (A).

Figure 2-21. Video Feedthrough T est Setup
Agilent 8360 P erformance T ests
Pulse Modulation Video Feedthrough

Pulse Generator HP/Agilent 8116A
Digitizing Oscilloscop e HP/Agilent 54111D
Preampli er HP/Agilent 8447F
130 MHz Filter K & L Micro w av e
500 MHz Low P ass Filter Mini-Circuits SLP-550
3.7 GHz Filter P/N 9135-0191
A tten uator 6 dB HP/Agilent 8493C Option 006
A tten uator 20 dB HP/Agilent 8493C Option 020
2-72 P erformance T ests
Pulse Modulation Video Feedthrough
Equipment

The pulse generator output lev el can b e a ected b y loading. Lea v e the pulse
generator connected to the oscilloscop e's c hannel 2 (through the BNC tee) as
y ou set the output lev el.
Channel 1:
Channel 2:
Timebase:
Delay Reference At center
Trigger:
Trigger Source Channel 1
Trigger Level 0.05 V
Display:
Number of Averages 16
and ampli er (B).
Channel 1:
Volts/Division 0.2 V
Offset 0.3 V
Agilent 8360 P erformance T ests
Pulse Modulation Video Feedthrough

Trigger:
Trigger Source Channel 1
Trigger Level 0.4 V
system Vp (from step 9)
Lo w Band Video Feedthrough
Disr e gar d if b eyond your synthesizer's c ap ability.
ampli er (D).
First CW frequency in T able 2-21
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ALC
Leveling Mode Search .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MOD
Pulse On/Off Extrnl (asterisk on)
or, for syn thesizers with Option 002, instead press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MOD
T able 2-21. Lo w Band Video Feedthrough Frequencies
2-74 P erformance T ests
Pulse Modulation Video Feedthrough

Channel 1:
Volts/Division 0.01 V
Channel 2:
Timebase:
Delay Reference At center
Trigger:
Trigger Source Channel 2
Trigger Level 1 V
Display:
Number of averages 64
(See Figure 2-22.) Video feedthrough (Vp) = Vp
Agilent 8360 P erformance T ests
Pulse Modulation Video Feedthrough


Figure 2-22. Measuring Maximum V oltage from the Baseline Video Feedthrough
carrier v oltage from T able 2-22 calculate the video feedthrough as a p ercent of carrier
p ow er:
Video feedthrough (%) = [(video Vp/carrier Vp)x100]/video gain
T able 2-22. P o w er Lev el Conv ersions to V olts P eak
Sp eci ed Po w er
Sp eci ed Po w er
(V)
0 0.316
10.5 1.059
0.5 0.335
11.0 1.122
1.0 0.355
11.5 1.189
1.5 0.376
12.0 1.259
2.0 0.398
12.5 1.334
2.5 0.422
13.0 1.413
3.0 0.447
13.5 1.496
3.5 0.473
14.0 1.585
4.0 0.501
14.5 1.679
4.5 0.531
15.0 1.778
5.0 0.562
15.5 1.884
5.5 0.596
16.0 1.995
6.0 0.631
16.5 2.113
6.5 0.668
17.0 2.239
7.0 0.708
17.5 2.371
7.5 0.750
18.0 2.512
8.0 0.794
18.5 2.661
8.5 0.841
19.0 2.819
9.0 0.891
19.5 2.985
9.5 0.944
20.0 2.985
10.0 1.000
2-76 P erformance T ests
Pulse Modulation Video Feedthrough

P eak Carrier V oltage = 10
rep eat steps 16 through 18 for all of the CW frequencies in T able 2-21 .
Video Feedthrough > 2 GHz
16 through 17 (the oscilloscop e and pulse generator are already set up) for the CW
frequencies in T able 2-23 . Do not exceed the maxim um frequency of y our syn thesizer.
Record the video feedthrough in T able 2-23 .
T able 2-23. Millimeter Band Video Feedthrough Frequencies
In Case of Difficulty
Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series Swept
CW Gener ator T r oublesho oting Guide .
Agilent 8360 P erformance T ests
Pulse Modulation Video Feedthrough

(F or B-Series Only)
This pro cedure measures the AM accuracy of the syn thesizer. The lo cal oscillator and mixer
are used to do wncon v ert the syn thesizer frequency to the frequency range of the measuring
receiv er (an HP/Agilent 11793A Micro w av e Con v erter can b e substituted for the mixer).
F or measuremen ts ab ov e 20 GHz RF output frequency , a sp ectrum analyzer with external
mixer are used to do wncon v ert to a 321.4 MHz RF. The function generator provides AM
mo dulation. The function generator signal lev el is adjusted for 30% AM as measured b y the
measuring receiv er. The function generator signal lev el is c hec k ed for accuracy . A DVM may
b e necessary to measure the function generator output.
up for at least one hour.
Lo w Band < 2 GHz
Disr e gar d if b eyond your synthesizer's c ap ability.

Figure 2-23. AM Accuracy T est Setup Lo w Band
2-78 P erformance T ests
Description and Procedure · AM Accuracy

F unction Generator HP/Agilent 3325A
Digital V oltmeter HP/Agilent 3456A
Measuring Receiv er HP/Agilent 8902A
A tten uator 10 dB HP/Agielnt 8490D Option 010 (83640B/50B)
A tten uator 10 dB HP/Agilent 8493C Option 010 (83620B/23B/30B)
POWER LEVEL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AM BW Cal Always (asterisk on)
Set the amplitude to 600 mV
) and v erify this amplitude with the DVM.
x 1.414 x 2 = V
).
MOD
AM On/Off 100%/V (asterisk on)
F or instrumen ts with Option 002, instead press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AM On/Off Ext (asterisk on)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AM Type 100%/V (asterisk on)
For Instruments with Option 002 only .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AM On/Off Int (asterisk on)
Internal AM Rate
Internal AM Depth
Agilent 8360 P erformance T ests

High Band > 2 GHz and < 20 GHz

Figure 2-24. AM Accuracy T est Setup < 20 GHz
F unction Generator HP/Agilent 3325A
Digital V oltmeter HP/Agilent 3456A
(Syn thesized Sw eeper) 83620A/B
Measuring Receiv er HP/Agilent 8902A
A tten uator 10 dB HP/Agilent 8490D Option 010
A tten uator 10 dB HP/Agilent 8493C Option 010
POWER LEVEL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
USER CAL
AM BW Cal Always (asterisk on)
2-80 P erformance T ests

Set the amplitude to 600 mV p-p and v erify this amplitude with the DVM. Note that the
DVM measures in V rms (V rms x 1.414 x 2 = V p-p).
MOD
AM On/Off 100%/V (asterisk on)
F or instrumen ts with Option 002, instead press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AM On/Off Ext (asterisk on)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AM Type 100%/V (asterisk on)
For Instruments with Option 002 only .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AM On/Off Ext (asterisk o )
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AM On/Off Int (asterisk on)
Internal AM Rate
Internal AM Depth
High Band > 20 GHz { Doubler Engaged
Disr e gar d if b eyond your synthesizer's c ap ability.
Agilent 8360 P erformance T ests


Figure 2-25. AM Accuracy T est Setup > 20 GHz
Sp ectrum Analyzer HP/Agilent 8566B
Micro w av e Ampli er HP/Agilent 11975A
Measuring Receiv er HP/Agilent 8902A
Harmonic Mixer P/N 0955-0307
F unction Generator HP/Agilent 3325A
A tten uator 10 dB HP/Agilent 8490D Option 010
POWER LEVEL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MOD
AM On/Off 100%/V (asterisk o )
F or instrumen ts with Option 002, instead press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MOD
AM On/Off Ext (asterisk on)
AM Type 100%/V (asterisk on)
AM On/Off 100%/V (asterisk on)
2-82 P erformance T ests

F or instrumen ts with Option 002, instead press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AM On/Off Ext (asterisk on)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AM Type 100%/V (asterisk on)
For Instruments with Option 002 only:.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AM On/Off Ext (asterisk o )
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AM On/Off Int (asterisk on)
Internal AM Rate
Internal AM Depth
Mo dulator O set and Gain
In Case of Difficulty
(the factory-set v alue is 50 ). If the AM input is set for 2 k , and the function generator
requires a 50 system, use a DVM to measure the AM input signal lev el while it is
connected to the syn thesizer.
Swept CW Gener ator T r oublesho oting Guide .
Agilent 8360 P erformance T ests

(F or B-Series Only)
This test v eri es that the p eak RF output p ow er do es not v ary more than a sp eci ed amoun t
ov er the sp eci ed amplitude mo dulation rate frequency range. The sp ectrum analyzer with
the tracking generator op erates as a net w ork analyzer to measure the atness. Flatness
errors asso ciated with the tracking generator and sp ectrum analyzer are calibrated out of the
measuremen t.
connected to the 50 ohm input of the sp ectrum analyzer and to the DVM. Press
and let the equipmen t w arm up for at least one hour.

Figure 2-26. AM Bandwidth T est Setup
With T rac king Generator HP/Agilent 3585A/B
Crystal Detector HP/Agilent 33330E (83640B/50B)
Crystal Detector HP/Agilent 33330D (83620B/22B/23B/24B/30B)
Digital V oltmeter HP/Agilent 3456A
2-84 P erformance T ests

, using the RPG
knob, to p osition the trace (dot) on the cen ter horizontal graticule.
trace A in to trace B memory . Then display A minus B (turning o trace B). A straight
line should b e display ed.
The sp ectrum analyzer is no w calibrated.
syn thesizer's RF output through the crystal detector to the sp ectrum analyzer's 50 ohm
input.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AM BW CAL Always (asterisk on)
AM CAL MENU
CW
First syn thesizer frequency in T able 2-24
Set to 5 dB b elow sp eci ed maximum lev eled p ow er
AM On/Off 100%/V (asterisk on)
or for syn thesizers with Option 002, press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
T able 2-24. Synthesizer Frequencies
1 Disregard if b ey ond your synthesizer's
capability .
set the p eak of the trace on the cen ter graticule. If an ov er range o ccurs, turn the RPG
knob in the opp osite direction.
The display no w sho ws the syn thesizer AM atness from 5 kHz to 200 kHz.
frequency of the 3 dB bandwidth in T able 2-24. If the 3 dB bandwidth p oint is o the
display , record \ > 200 kHz" in T able 2-24.
Agilent 8360 P erformance T ests

and rep eat steps 7 and 8 for the remaining
syn thesizer frequencies in T able 2-24.
In Case of Difficulty
Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series
Swept CW Gener ator T r oublesho oting Guide .
2-86 P erformance T ests
Related Adjustments · None · AM Bandwidth

(F or B-Series Only)
In this pro cedure, a function generator is used to provide a DC v oltage to the syn thesizer's
AM input. The syn thesizer is set to 0 dBm output p ow er. AM is enabled and a DC v oltage
is applied. The DC v oltage is set for maximum AM without an ov ermo dulation indication to
drive the syn thesizer's mo dulator to the edge of distortion. The output p ow er is measured
with a p ow er meter.
A measuremen t is made in 1 GHz steps across the syn thesizer's frequency range. This
equipmen t lets y ou measure do wn to 30 dBm. If y ou reac h this lev el b efore the syn thesizer
ov ermo dulates, the p ow er meter will indicate an under range measuremen t.
This pro cedure is rep eated using the deep AM function in the searc h ALC mo de. A sp ectrum
analyzer is used in place of the p ow er meter to make the deep AM measuremen t. This
equipmen t allo ws y ou to measure do wn to approximately 65 dBm. If y our syn thesizer
do es not ov ermo dulate b y this p oint, and y ou wish to measure the p ow er lev el just b efore
ov ermo dulation, use a p ow er supply to increase the DC v oltage supply to the AM input.
Do not exceed 15 V DC to the AM input or damage will o ccur.

Figure 2-27. AM Dynamic Range T est Setup
Agilent 8360 P erformance T ests
Description and Procedure · AM Dynamic Range

Sp ectrum Analyzer HP/Agilent 8566B
Digital V oltmeter HP/Agilent 3456A
F unction Generator HP/Agilent 3325A
P ow er Meter HP/Agilent 436A/7A/8A
P ow er Sensor HP/Agilent 8487A (83640B/50B)
P ow er Sensor HP/Agilent 8485A (83620B/22B/23B/24B/30B)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
USER CAL
Peak RF Always (asterisk on)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRIOR
AM BW Cal Always (asterisk on)
syn thesizer's RF OUTPUT.
(Set to 2 GHz for syn thesizers with low est start frequency of 2 GHz)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Up/Dn Size CW
PO WER
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Uncoupl Atten (Option 001 only)
MOD
V erify that all mo dulation is o . (No asterisks next to k ey lab els.)
F or syn thesizers with Option 002, if mo dulation is on, an \AM," \FM," or \PLS" message
will b e display ed.
Normal AM Dynamic Range
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AM On/Off 10dB/V (asterisk on)
or, for syn thesizers with Option 002, press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MOD
AM On/Off Ext (asterisk on)
AM Type 10 dB/V (asterisk on)
Then increase the signal un til the O VERMOD message turns o . (The more negative the
signal applied to the mo dulator is, the more the RF output level decreases.)
p ow er meter reads under range, record < 30 dB.
2-88 P erformance T ests

T able 2-25. P o w er Meter Readings
1 GHz 12 GHz
2 GHz 13 GHz
3 GHz 14 GHz
4 GHz 15 GHz
5 GHz 16 GHz
6 GHz 17 GHz
7 GHz 18 GHz
8 GHz 19 GHz
9 GHz 20 GHz
10 GHz 21 GHz
11 GHz 22 GHz
1 83630B/40B/50B only .
factor % as required.
Deep AM Dynamic Range
sho wn in Figure 2-27.
(Set to 2 GHz for syn thesizers with low est start frequency of 2 GHz)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AM On/Off 10 dB/V (asterisk o )
or for syn thesizers with Option 002, press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AM On/Off Int (asterisk o )
AM On/Off 100%/V (asterisk on)
Deep AM (asterisk on)
or for syn thesizers with Option 002, press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AM Type 100%/V (asterisk o )
Deep AM (asterisk on)
Agilent 8360 P erformance T ests

Make the follo wing measuremen ts as quic kly as p ossible since the signal
amplitude is sub ject to drift when y ou use deep AM at low lev els. If necessary ,
turning AM o and on again will reset the p ow er lev el.
message on the syn thesizer or un til the signal drops in to the noise lev el on the sp ectrum
analyzer. Decrease the function generator output just un til the O VERMOD message
disapp ears.
T able 2-26. RF Signal Lev el
1 GHz 12 GHz
2 GHz 13 GHz
3 GHz 14 GHz
4 GHz 15 GHz
5 GHz 16 GHz
6 GHz 17 GHz
7 GHz 18 GHz
8 GHz 19 GHz
9 GHz 20 GHz
10 GHz 21 GHz
11 GHz 22 GHz
1 83630B/40B/50B only .
AM On/Off 10dB/V (asterisk o )
or for syn thesizers with Option 002, press:
AM Type 10 dB/V (asterisk o )
T urn mo dulation bac k on and rep eat steps 19 through 21 to the highest frequency from
T able 2-26 that applies to y our syn thesizer.
a. F or the 83630B , record w orst case v alues for frequencies < 20 GHz and for frequencies
20 GHz.
b. F or the 83640B /50B, record w orst case v alues for frequencies 20 GHz and for
frequencies > 20 GHz.
c. F or all other instrumen ts, record the w orst case v alue for all frequencies tested.
2-90 P erformance T ests
AM Dynamic Range

Mo dulator O set and Gain
Mo dulation Generator Adjustmen t (Option 002)
In Case of Difficulty
Swept CW Gener ator T r oublesho oting Guide .
Agilent 8360 P erformance T ests
Related Adjustments · AM Dynamic Range

(F or B-Series Only)
Use this pro cedure to measure the FM accuracy of the syn thesizer. With the syn thesizer set
to an FM sensitivit y of 10 MHz/V, and the function generator providing an FM mo dulation
index of 2.404, the RF output should b e a n ull. The function generator amplitude is adjusted
for the Bessel n ull and the di erence b et w een signals is recorded.
Since the mo dulation frequency for external FM is 1 MHz, this pro cedure
relies on the v oltage-setting accuracy of the function generator (the frequency
is to o high for a DVM, and an oscilloscop e is not accurate enough.)
w arm up for at least one hour.

Figure 2-28. FM Accuracy T est Setup
Sp ectrum Analyzer HP/Agilent 8566B
F unction Generator HP/Agilent 3325A
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
USER CAL
Peak RF Always (asterisk on)
First syn thesizer frequency in T able 2-27
Set the sp eci ed maximum lev eled p ow er
2-92 P erformance T ests

T able 2-27. Synthesizer Frequencies
Syn thesizer F requencies
1 Disregard if b eyond your synthesizer's capability .
MOD
FM On/Off 100 kHz (asterisk on)
or for syn thesizers with Option 002, press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FM On/Off Ext (asterisk on)
sp ectrum analyzer.
Error (%) = [(480.8 mV p-p New Amplitude)/480.8 mV p-p] x 100.
FM On/Off 100 kHz (asterisk o )
or for syn thesizers with Option 002, press:
FM On/Off Ext (asterisk o )
T able 2-27.
Agilent 8360 P erformance T ests

Internal FM (Option 002 only)
First syn thesizer frequency in T able 2-27
NNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MOD
FM On/Off Ext (asterisk o )
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MOD
FM On/Off Int (asterisk on)
Internal FM Rate
Internal FM Dev
2.404
carrier on the sp ectrum analyzer.
Error (%) = [(2.404 MHz New FM Deviation)/2.404 MHz] x 100.
FM On/Off Int (asterisk o )
T able 2-27.
Mo dulation Generator (Option 002)
Mo dulation Generator Flatness (Option 002)
In case of Difficulty
Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series
Swept CW Gener ator T r oublesho oting Guide .
2-94 P erformance T ests

(F or B-Series Only)
Use this pro cedure to v erify that the RF output p ow er do es not v ary more than a sp eci ed
amoun t ov er the sp eci ed frequency mo dulation rate frequency range. A low frequency
sp ectrum analyzer with a tracking generator op erates as a net w ork analyzer to measure the
FM atness. Flatness errors asso ciated with the tracking generator and sp ectrum analyzer are
calibrated out of the measuremen t. Flatness errors due to the p ow er splitter and microw a v e
ampli er are minimized b y external lev eling at the ampli er. The dela y line discriminator is
used to demo dulate the FM from the RF carrier. It also provides an input to the sp ectrum
analyzer that is equiv alen t to the tracking generator frequency and prop ortional in p ow er to
the syn thesizer RF output.
The dela y line discriminator is formed b y driving a mixer with a mo dulated carrier in to
the LO p ort, and a dela y ed carrier in to the RF p ort. The dela y is pro duced with a cable of
approximately 3 feet in length. With this setup, the mixer resp onds to phase di erences
b et w een the tw o inputs. The cable has a constant time dela y and, as the carrier frequency is
frequency-mo dulated, a linear phase shift o ccurs b et w een the mixer inputs. Since the mixer
inputs are in quadrature, the mixer op erates as a phase detector that amplitude tracks the
RF input (syn thesizer p ow er) and has an output frequency corresp onding to the FM rate
(tracking generator frequency).
Agilent 8360 P erformance T ests
Description and Procedure · FM Bandwidth

connected to the c hannel 1 input of the oscilloscop e and to the FM input of the
syn thesizer (A). A DVM can b e used in place of the oscilloscop e for rms measuremen ts.
Preset all the equipmen t and let them w arm up at least one hour.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Peak RF Always (asterisk on)
Set the sp ectrum analyzer tracking generator to 100 mV p-p (35 mV rms). Cho ose the
alternate detector if y our tracking generator has an alternate detector that improv es the
low-end atness.
through the 10 dB attenuator (B). On the sp ectrum analyzer, set:
2-96 P erformance T ests


Figure 2-29. FM Bandwidth T est Setup
Agilent 8360 P erformance T ests
FM Bandwidth

Sp ectrum Analyzer with
T rac king Generator
Digitizing Oscilloscop e HP/Agilent 54111D
Digital V oltmeter HP/Agilent 3456A
Micro w av e Ampli er HP/Agilent 8348A
P ow er Splitter HP/Agilent 11667C (83640B/50B)
P ow er Splitter HP/Agilent 11667B
A tten uator 10 dB (2) HP/Agilent 8490D Option 010
A tten uator 10 dB (2) HP/Agilent 8493C Option 010
P/N 08503-20038
Discriminator (cable 3 ft.)
RPG knob, to p osition the trace on the cen ter horizontal graticule.
minus B (turning o trace B). A straight line should b e display ed.
mixer's IF output to the DVM's V OL TS HI/LO input (C).
Set the DVM to measure DC v olts.
First syn thesizer frequency in T able 2-28
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ALC
POWER LEVEL
Set the p ow er so that the microw av e ampli er displays an output of 16
dBm.
T able 2-28. Synthesizer Frequencies
Syn thesizer F requencies
1 Disregard if b eyond your synthesizer's capability .
2 83630B/40B/50B only .
2-98 P erformance T ests

and use the left arrow k ey to p osition the cursor in the 1
MHz p osition (one digit left of the decimal p oint). Use the rotary knob or step k eys to
adjust the CW frequency for 0 V DC on the DVM.
DC FM Flatness (Unlock ed)
FM On/Off DC (asterisk on)
Change the reference lev el to set the start of the trace or the highest p eak on the cen ter
graticule.
the sp ectrum analyzer markers to nd the 3 dB bandwidth. Record this frequency in
T able 2-28. If the total v ariation is less than 3 dB then record \ > 10 MHz" in T able 2-28.
In Case of Difficulty
Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series
Swept CW Gener ator T r oublesho oting Guide .
Agilent 8360 P erformance T ests

(F or B-Series Only)
In this pro cedure, the syn thesizer's RF output is monitored directly on a sp ectrum analyzer
for low FM rates. The FM amplitude is increased un til the maximum deviation exceeds the
sp eci cation.
F or higher FM rates, the FM rate is set so that a Bessel n ull o ccurs when the RF p eak
deviation equals the sp eci cation. The mo dulation amplitude is increased in order to pass
through the correct Bessel n ull for the FM rate used and the p eak deviation sp eci ed.
w arm up for at least one hour.

Figure 2-30. Maximum FM Deviation T est Setup
Sp ectrum Analyzer HP/Agilent 8566B
F unction Generator HP/Agilent 3325A
F unction Generator HP/Agilent 8111A or HP/Agilent 8116A
DC FM (100 Hz) (Using HP/Agilent 8111A or HP/Agilent 8116A)
Syn thesizer mo dels without 1 GHz capabilit y should use 5 GHz settings
instead of 1 GHz, as indicated in paren thesis.
2-100 P erformance T ests

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Up/Dn Size CW
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
USER CAL
Peak RF Always (asterisk on)
the sp ectrum analyzer to maximum hold.
MOD
FM On/Off DC (asterisk on)
or for syn thesizers with Option 002, press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FM On/Off Ext (asterisk on)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FM Coupling DC (asterisk on)
The function generator m ust b e capable of supplying > 16 Vp-p in to 50 to
obtain sucien t FM deviation for this step.
eigh t divisions on the sp ectrum analyzer. In T able 2-29 record P ASS if the signal exceeds
eigh t divisions with no unlo c k or ov ermo d message on the syn thesizer. Record F AIL if
not.
T able 2-29. DC FM T est P ass-F ail Results
1 Disregard if b eyond your synthesizer's capability .
FM On/Off DC (asterisk o )
or for syn thesizers with Option 002, press:
FM On/Off Ext (asterisk o )
frequencies in T able 2-29 .
Agilent 8360 P erformance T ests

If not, circle F AIL on the test record.
A C FM (at 100 kHz) (Using HP/Agilent 3325A)
Maxim um FM deviation for a 100 kHz rate is limited to a mo dulation index of
5 times n, where n = frequency band. The mo dulation index is equiv alen t to
deviation/rate.
Syn thesizer mo dels without 1 GHz capabilit y should use 5 GHz settings, as
indicated in paren thesis, instead of 1 GHz.
Clear all data from the display.
FM On/Off 100 kHz (asterisk on)
or, for syn thesizers with Option 002, press:
NNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FM On/Off Ext (asterisk on)
This activ ates FM sensitivity . Use the do wn arrow k ey to set FM sensitivity to 1.00
MHz/V.
k ey to do an amplitude calibration. Use the arrow k eys to adjust the
amplitude to ac hiev e a n ull of the 1st sideband (Sideband Nulled).
The sidebands may not resp ond to increases and decreases of the FM input
v oltage symmetrically . In this case, take the av erage of the v oltages that cause
the low er sideband and upp er sideband to n ull. If the sidebands n ull at more
than one v oltage lev el, use the low er v oltage v alue as the n ull v oltage.
Calculated Null V oltage
Actual FM Sensitivity =
Actual Null V oltage
Record this v alue in the space provided in T able 2-30.
2-102 P erformance T ests

syn thesizer displays an UNLOCK or O VERMOD message. Decrease the FM Input
V oltage just un til the UNLOCK or O VERMOD message turns o .
Max FM Deviation = (Actual FM Sensitivit y)[(1.414)(Max FM V oltage)]
Remem b er to use the set FM sensitivit y as a m ultiplier when calculating the
actual FM sensitivit y .
T able 2-30. FM Deviation Frequencies and Settings

1 Disregard if b eyond your synthesizer's capabilit y .
Agilent 8360 P erformance T ests
Maximum FM Deviation

Mo dulation Generator Flatness
In Case of Difficulty
for the CW frequency in question.
Swept CW Gener ator T r oublesho oting Guide .
2-104 P erformance T ests

(F or B-Series Only)
This p erformance test uses an oscilloscop e to measure the pulse width of the output of the
mo dulation generator (not the pulse width of the RF output).
up for at least one hour.

Figure 2-31. Pulse Accuracy T est Setup
Digitizing Oscilloscop e HP/Agilent 54111D
Oscilloscop e Prob es HP/Agilent 10437A
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Pulse On/Off Int (asterisk on)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Internal Pulse Width
Internal Pulse Period
Agilent 8360 P erformance T ests

Channel 1:
Timebase:
Trigger:
Trig Src Chan 1
Trigger Level 1.2V
Display:
Display Time 0.2 s
test record.
Internal Pulse Width
test record.
In Case of Difficulty
Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series
Swept CW Gener ator T r oublesho oting Guide .
2-106 P erformance T ests

(F or B-Series Only)
The mo dulation meter measures and displays the v alue of the externally generated amplitude
or frequency mo dulation. This pro cedure determines the accuracy of the mo dulation meter.
w arm up for at least one hour.

Figure 2-32. Modulation Meter T est Setup
F unction Generator HP/Agilent 3325A
Digital V oltmeter HP/Agilent 3456A
N CYC INT
STORE
N DIG DISP
.
AM meter function. Press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AM On/Off Ext (asterisk on).
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Meter On/Off AM (asterisk on).
adjust the function generator for a DVM reading of 0.6363 A C V rms.
1 MHz/V, and turn on the FM meter function. Press:
NNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AM On/Off Ext (asterisk o )
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRIOR
FM On/Off Ext (asterisk on)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRIOR
Meter On/Off FM (asterisk on)
Agilent 8360 P erformance T ests

adjust the function generator amplitude for a DVM reading of 3.535 A C V rms.
In Case of Difficulty
Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series
Swept CW Gener ator T r oublesho oting Guide .
2-108 P erformance T ests
Related Adjustments · None · Modulation Meter

T able 2-33. T est Record for Agilent 83620B and 83622B (1 of 8)
Mo del Am bien t temp erature
Serial Num b er Relative humidit y %
Options Line frequency Hz (nominal)
Firm w are Revision
Sp ecial Notes:
Agilent 8360 P erformance T ests
Agilent 83620B/22B T est Record

T able 2-33. T est Record for 83620B and 83622B (2 of 8)
Mo del Rep ort Num b er Date
T est Equipmen t Used Mo del Numb er T race Numb er Cal Due Date
2-110 P erformance T ests
Agilent 83620B/22B T est Record

T able 2-33. T est Record for 83620B and 83622B (3 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertain t y
< 2 GHz 0.6 dB +0.6 dB 0.19 dB
2 and 20 GHz 0.7 dB +0.7 dB 0.35 dB
< 2 GHz 1.2 dB +1.2 dB 0.21 dB
2 and 20 GHz 1.3 dB +1.3 dB 0.46 dB
2 and 20 GHz 1.2 dB 0.25 dB
Agilent 8360 P erformance T ests
Agilent 83620B/22B T est Record

T able 2-33. T est Record for 83620B and 83622B (4 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertaint y
100 MHz Fixed 60 dBc 1 dB
100 MHz O set 60 dBc 1 dB
LO F eedthrough 60 dBc 1 dB
2-112 P erformance T ests
Agilent 83620B/22B T est Record

T able 2-33. T est Record for 83620B and 83622B (5 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertain t y
120 Hz Spur 55 dBc 1 dB
180 Hz Spur 55 dBc 1 dB
240 Hz Spur 55 dBc 1 dB
120 Hz Spur 55 dBc 1 dB
180 Hz Spur 55 dBc 1 dB
240 Hz Spur 55 dBc 1 dB
100 Hz to < 1 kHz 70 dBc 2 dB
1 kHz to < 10 kHz 78 dBc 2 dB
10 kHz to < 100 kHz 86 dBc 2 dB
100 kHz 107 dBc 2 dB
100 Hz to < 1 kHz 70 dBc 2 dB
1 kHz to < 10 kHz 78 dBc 2 dB
10 kHz to < 100 kHz 86 dBc 2 dB
100 kHz 107 dBc 2 dB
100 Hz to < 1 kHz 60 dBc 2 dB
1 kHz to < 10 kHz 68 dBc 2 dB
10 kHz to < 100 kHz 76 dBc 2 dB
100 kHz 97 dBc 2 dB
1 3 dB if the Invertron is not used.
Agilent 8360 P erformance T ests
Agilent 83620B/22B T est Record

T able 2-33. T est Record for 83620B and 83622B (6 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertaint y
2-114 P erformance T ests
Agilent 83620B/22B T est Record

T able 2-33. T est Record for 83620B and 83622B (7 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertain t y
Agilent 8360 P erformance T ests
Agilent 83620B/22B T est Record

T able 2-33. T est Record for 83620B and 83622B (8 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertaint y
> 75 MHz Deviation P ass/F ail 2.3 MHz
2-116 P erformance T ests
Agilent 83620B/22B T est Record

T able 2-34. T est Record for Agilent 83623B and 83624B (1 of 8)
Mo del Am bien t temp erature
Serial Num b er Relative humidit y %
Options Line frequency Hz (nominal)
Firm w are Revision
Sp ecial Notes:
Agilent 8360 P erformance T ests
Agilent 83623B/24B T est Record

T able 2-34. T est Record for 83623B and 83624B (2 of 8)
Mo del Rep ort Num b er Date
T est Equipmen t Used Mo del Numb er T race Numb er Cal Due Date
2-118 P erformance T ests
Agilent 83623B/24B T est Record

T able 2-34. T est Record for 83623B and 83624B (3 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertain t y
< 2 GHz 0.6 dB +0.6 dB 0.19 dB
2 and 20 GHz 0.7 dB +0.7 0.35 dB
< 2 GHz 1.2 dB +1.2 dB 0.21 dB
2 and 20 GHz 1.3 dB +1.3 dB 0.46 dB
2 and 20 GHz 1.2 dB 0.25 dB
Standard or Opt. 006 +17 dBm 1.8 dB
Opt. 001 or Opts. 001 and 006 +15.5 dBm 1.8 dB
Agilent 8360 P erformance T ests
Agilent 83623B/24B T est Record

T able 2-34. T est Record for 83623B and 83624B (4 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertaint y
Opts. 001 and 006 +15.5 dBm 1.8 dB
2-120 P erformance T ests
Agilent 83623B/24B T est Record

T able 2-34. T est Record for 83623B and 83624B (5 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertain t y
100 MHz Fixed 60 dBc 1 dB
100 MHz O set 60 dBc 1 dB
LO F eedthrough 60 dBc 1 dB
120 Hz Spur 55 dBc 1 dB
180 Hz Spur 55 dBc 1 dB
240 Hz Spur 55 dBc 1 dB
120 Hz Spur 55 dBc 1 dB
180 Hz Spur 55 dBc 1 dB
240 Hz Spur 55 dBc 1 dB
100 Hz to < 1 kHz 70 dBc 2 dB
1 kHz to < 10 kHz 78 dBc 2 dB
10 kHz to < 100 kHz 86 dBc 2 dB
100 kHz 107 dBc 2 dB
100 Hz to < 1 kHz 70 dBc 2 dB
1 kHz to < 10 kHz 78 dBc 2 dB
10 kHz to < 100 kHz 86 dBc 2 dB
100 kHz 107 dBc 2 dB
1 3 dB if the Invertron is not used.
Agilent 8360 P erformance T ests
Agilent 83623B/24B T est Record

T able 2-34. T est Record for 83623B and 83624B (6 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertaint y
100 Hz to < 1 kHz 60 dBc 2 dB
1 kHz to < 10 kHz 68 dBc 2 dB
10 kHz to < 100 kHz 76 dBc 2 dB
100 kHz 97 dBc 2 dB
2-122 P erformance T ests
Agilent 83623B/24B T est Record

T able 2-34. T est Record for 83623B and 83624B (7 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertain t y
Agilent 8360 P erformance T ests
Agilent 83623B/24B T est Record

T able 2-34. T est Record for 83623B and 83624B (8 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertaint y
> 75 MHz Deviation P ass/F ail 2.3 MHz
2-124 P erformance T ests
Agilent 83623B/24B T est Record

T able 2-35. T est Record for Agilent 83630B (1 of 8)
Mo del Am bien t temp erature
Serial Num b er Relative humidit y %
Options Line frequency Hz (nominal)
Firm w are Revision
Sp ecial Notes:
Agilent 8360 P erformance T ests
Agilent 83630B T est Record

T able 2-35. T est Record for 83630B (2 of 8)
Mo del Rep ort Num b er Date
T est Equipmen t Used Mo del Numb er T race Numb er Cal Due Date
2-126 P erformance T ests
Agilent 83630B T est Record

T able 2-35. T est Record for 83630B (3 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertain t y
< 2 GHz 0.6 dB +0.6 dB 0.19 dB
2 and 20 GHz 0.7 dB +0.7 dB 0.35 dB
> 20 and 26.5 GHz 0.9 dB +0.9 dB 0.38 dB
< 2 GHz 1.2 dB +1.2 dB 0.21 dB
2 and 20 GHz 1.3 dB +1.3 dB 0.46 dB
2 and 20 GHz 1.2 dB 0.25 dB
> 20 and 26.5 GHz 1.6 dB 0.27 dB
0.01 to 26.5 GHz +10.0 dBm 1.7 dB
0.01 to 20 GHz +13.0 dBm 1.7 dB
Agilent 8360 P erformance T ests
Agilent 83630B T est Record

T able 2-35. T est Record for 83630B (4 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertaint y
0.01 to 26.5 GHz +8.0 dBm 1.7 dB
0.01 to 20 GHz +11.5 dBm 1.7 dB
2-128 P erformance T ests
Agilent 83630B T est Record

T able 2-35. T est Record for 83630B (5 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertain t y
100 MHz Fixed 60 dBc 1 dB
100 MHz O set 60 dBc 1 dB
LO F eedthrough 60 dBc 1 dB
120 Hz Spur 55 dBc 1 dB
180 Hz Spur 55 dBc 1 dB
240 Hz Spur 55 dBc 1 dB
120 Hz Spur 55 dBc 1 dB
180 Hz Spur 55 dBc 1 dB
240 Hz Spur 55 dBc 1 dB
100 Hz to < 1 kHz 70 dBc 2 dB
1 kHz to < 10 kHz 78 dBc 2 dB
10 kHz to < 100 kHz 86 dBc 2 dB
100 kHz 107 dBc 2 dB
100 Hz to < 1 kHz 70 dBc 2 dB
1 kHz to < 10 kHz 78 dBc 2 dB
10 kHz to < 100 kHz 86 dBc 2 dB
100 kHz 107 dBc 2 dB
1 3 dB if the Invertron is not used.
Agilent 8360 P erformance T ests
Agilent 83630B T est Record

T able 2-35. T est Record for 83630B (6 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertaint y
100 Hz to < 1 kHz 60 dBc 2 dB
1 kHz to < 10 kHz 68 dBc 2 dB
10 kHz to < 100 kHz 76 dBc 2 dB
100 kHz 97 dBc 2 dB
2-130 P erformance T ests
Agilent 83630B T est Record

T able 2-35. T est Record for 83630B (7 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertain t y
Agilent 8360 P erformance T ests
Agilent 83630B T est Record

T able 2-35. T est Record for 83630B (8 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertaint y
> 75 MHz Deviation P ass/F ail 2.3 MHz
2-132 P erformance T ests
Agilent 83630B T est Record

T able 2-36. T est Record for Agilent 83640B (1 of 8)
Mo del Am bien t temp erature
Serial Num b er Relative humidit y %
Options Line frequency Hz (nominal)
Firm w are Revision
Sp ecial Notes:
Agilent 8360 P erformance T ests
Agilent 83640B T est Record

T able 2-36. T est Record for 83640B (2 of 8)
Mo del Rep ort Num b er Date
T est Equipmen t Used Mo del Numb er T race Numb er Cal Due Date
2-134 P erformance T ests
Agilent 83640B T est Record

T able 2-36. T est Record for 83640B (3 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertain t y
< 2 GHz 0.6 dB 0.6 dB 0.22 dB
2 and 20 GHz 0.7 dB 0.7 dB 0.35 dB
> 20 and 40 GHz 0.9 dB 0.9 dB 0.38 dB
2 and 20 GHz 1.2 dB 0.25 dB
> 20 and 40 GHz 1.8 dB 0.27 dB
0.01 to 40 GHz +6.0 dBm 1.4 dB
0.01 to < 26.5 GHz +10.0 dBm 1.8 dB
0.01 to < 20 GHz +10.0 dBm 1.8 dB
Agilent 8360 P erformance T ests
Agilent 83640B T est Record

T able 2-36. T est Record for the 83640B (4 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertaint y
0.01 to 40 GHz +4 dBm 1.4 dB
0.01 to < 26.5 GHz +8 dBm 1.8 dB
0.01 to < 20 GHz +8.5 dBm 1.8 dB
100 MHz Fixed 60 dBc 1 dB
100 MHz O set 60 dBc 1 dB
LO F eedthrough 60 dBc 1 dB
2-136 P erformance T ests
Agilent 83640B T est Record

T able 2-36. T est Record for the 83640B (5 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertain t y
100 Hz to < 1 kHz 70 dBc 2 dB
1 kHz to < 10 kHz 78 dBc 2 dB
10 kHz to < 100 kHz 86 dBc 2 dB
100 kHz 107 dBc 2 dB
100 Hz to < 1 kHz 70 dBc 2 dB
1 kHz to < 10 kHz 78 dBc 2 dB
10 kHz to < 100 kHz 86 dBc 2 dB
100 kHz 107 dBc 2 dB
100 Hz to < 1 kHz 60 dBc 2 dB
1 kHz to < 10 kHz 68 dBc 2 dB
10 kHz to < 100 kHz 76 dBc 2 dB
100 kHz 97 dBc 2 dB
1 3 dB if the Invertron is not used.
Agilent 8360 P erformance T ests
Agilent 83640B T est Record

T able 2-36. T est Record for the 83640B (6 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertaint y
2-138 P erformance T ests
Agilent 83640B T est Record

T able 2-36. T est Record for the 83640B (7 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertain t y
Agilent 8360 P erformance T ests
Agilent 83640B T est Record

T able 2-36. T est Record for the 83640B (8 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertaint y
> 75 MHz Deviation P ass/F ail 2.3 MHz
2-140 P erformance T ests
Agilent 83640B T est Record

T able 2-37. T est Record for Agilent 83650B (1 of 8)
Mo del Am bien t temp erature
Serial Num b er Relative humidit y %
Options Line frequency Hz (nominal)
Firm w are Revision
Sp ecial Notes:
Agilent 8360 P erformance T ests
Agilent 83650B T est Record

T able 2-37. T est Record for 83650B (2 of 8)
Mo del Rep ort Num b er Date
T est Equipmen t Used Mo del Numb er T race Numb er Cal Due Date
2-142 P erformance T ests
Agilent 83650B T est Record

T able 2-37. T est Record for 83650B (3 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertain t y
< 2 GHz 0.6 dB 0.6 dB 0.22 dB
2 and 20 GHz 0.7 dB 0.7 dB 0.35 dB
> 20 and 40 GHz 0.9 dB 0.9 dB 0.38 dB
> 40 GHz 1.7 dB 1.7 dB 0.79 dB
2 and 20 GHz 1.2 dB 0.25 dB
> 20 and 40 GHz 1.6 dB 0.38 dB
0.01 to 50 GHz +2.5 dBm 2.3 dB
0.01 to < 40 GHz +5.0 dBm 1.3 dB
0.01 to < 26.5 GHz +10.0 dBm 1.7 dB
0.01 to < 20 GHz +10.0 dBm 1.7 dB
Agilent 8360 P erformance T ests
Agilent 83650B T est Record

T able 2-37. T est Record for 83650B (4 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertaint y
0.01 to 50 GHz 0 dBm 2.3 dB
0.01 to < 40 GHz +3.0 dBm 1.3 dB
0.01 to < 26.5 GHz +8.0 dBm 1.7 dB
0.01 to < 20 GHz +8.5 dBm 1.7 dB
2-144 P erformance T ests
Agilent 83650B T est Record

T able 2-37. T est Record for 83650B (5 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertain t y
100 MHz Fixed 60 dBc 1 dB
100 MHz O set 60 dBc 1 dB
LO F eedthrough 60 dBc 1 dB
120 Hz Spur 55 dBc 1 dB
180 Hz Spur 55 dBc 1 dB
240 Hz Spur 55 dBc 1 dB
120 Hz Spur 55 dBc 1 dB
180 Hz Spur 55 dBc 1 dB
240 Hz Spur 55 dBc 1 dB
100 Hz to < 1 kHz 70 dBc 2 dB
1 kHz to < 10 kHz 78 dBc 2 dB
10 kHz to < 100 kHz 86 dBc 2 dB
100 kHz 107 dBc 2 dB
100 Hz to < 1 kHz 70 dBc 2 dB
1 kHz to < 10 kHz 78 dBc 2 dB
10 kHz to < 100 kHz 86 dBc 2 dB
100 kHz 107 dBc 2 dB
1 3 dB if the Invertron is not used.
Agilent 8360 P erformance T ests
Agilent 83650B T est Record

T able 2-37. T est Record for 83650B (6 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertaint y
100 Hz to < 1 kHz 60 dBc 2 dB
1 kHz to < 10 kHz 68 dBc 2 dB
10 kHz to < 100 kHz 76 dBc 2 dB
100 kHz 97 dBc 2 dB
2-146 P erformance T ests
Agilent 83650B T est Record

T able 2-37. T est Record for 83650B (7 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertain t y
Agilent 8360 P erformance T ests
Agilent 83650B T est Record

T able 2-37. T est Record for 83650B (8 of 8)

No. T est Description Sp ec. Results Sp ec. Uncertaint y
> 75 MHz Deviation P ass/F ail 2.3 MHz
2-148 P erformance T ests
Agilent 83650B T est Record

T able 2-34. T est Record for Agilent 83623L (1 of 5)
Mo del Am bien t temp erature
Serial Num b er Relative humidit y %
Options Line frequency Hz (nominal)
Firm w are Revision
Sp ecial Notes:
Agilent 8360 P erformance T ests
Agilent 83623L T est Record

T able 2-34. T est Record for 83623L (2 of 5)
Mo del Rep ort Num b er Date
T est Equipmen t Used Mo del Numb er T race Numb er Cal Due Date
2-150 P erformance T ests
Agilent 83623L T est Record

T able 2-34. T est Record for 83623L (3 of 5)

No. T est Description Sp ec. Results Sp ec. Uncertain t y
< 2 GHz 0.6 dB +0.6 dB 0.19 dB
2 and 20 GHz 0.7 dB +0.7 0.35 dB
< 2 GHz 1.2 dB +1.2 dB 0.21 dB
2 and 20 GHz 1.3 dB +1.3 dB 0.46 dB
2 and 20 GHz 1.2 dB 0.25 dB
Agilent 8360 P erformance T ests
Agilent 83623L T est Record

T able 2-34. T est Record for 83623L (4 of 5)

No. T est Description Sp ec. Results Sp ec. Uncertaint y
100 MHz Fixed 60 dBc 1 dB
100 MHz O set 60 dBc 1 dB
LO F eedthrough 60 dBc 1 dB
2-152 P erformance T ests
Agilent 83623L T est Record

T able 2-34. T est Record for 83623L (5 of 5)

No. T est Description Sp ec. Results Sp ec. Uncertain t y
120 Hz Spur 55 dBc 1 dB
180 Hz Spur 55 dBc 1 dB
240 Hz Spur 55 dBc 1 dB
120 Hz Spur 55 dBc 1 dB
180 Hz Spur 55 dBc 1 dB
240 Hz Spur 55 dBc 1 dB
100 Hz to < 1 kHz 70 dBc 2 dB
1 kHz to < 10 kHz 78 dBc 2 dB
10 kHz to < 100 kHz 86 dBc 2 dB
100 kHz 107 dBc 2 dB
100 Hz to < 1 kHz 70 dBc 2 dB
1 kHz to < 10 kHz 78 dBc 2 dB
10 kHz to < 100 kHz 86 dBc 2 dB
100 kHz 107 dBc 2 dB
100 Hz to < 1 kHz 60 dBc 2 dB
1 kHz to < 10 kHz 68 dBc 2 dB
10 kHz to < 100 kHz 76 dBc 2 dB
100 kHz 97 dBc 2 dB
1 3 dB if the Invertron is not used.
Agilent 8360 P erformance T ests
Agilent 83623L T est Record

T able 2-35. T est Record for Agilent 83630L (1 of 5)
Mo del Am bien t temp erature
Serial Num b er Relative humidit y %
Options Line frequency Hz (nominal)
Firm w are Revision
Sp ecial Notes:
2-154 P erformance T ests
Agilent 83630L T est Record

T able 2-35. T est Record for 83630L (2 of 5)
Mo del Rep ort Num b er Date
T est Equipmen t Used Mo del Numb er T race Numb er Cal Due Date
Agilent 8360 P erformance T ests
Agilent 83630L T est Record

T able 2-35. T est Record for 83630L (3 of 5)

No. T est Description Sp ec. Results Sp ec. Uncertaint y
< 2 GHz 0.6 dB +0.6 dB 0.19 dB
2 and 20 GHz 0.7 dB +0.7 dB 0.35 dB
> 20 and 26.5 GHz 0.9 dB +0.9 dB 0.38 dB
< 2 GHz 1.2 dB +1.2 dB 0.21 dB
2 and 20 GHz 1.3 dB +1.3 dB 0.46 dB
2 and 20 GHz 1.2 dB 0.25 dB
> 20 and 26.5 GHz 1.6 dB 0.27 dB
0.01 to 26.5 GHz +10.0 dBm 1.7 dB
0.01 to 20 GHz +13.0 dBm 1.7 dB
2-156 P erformance T ests
Agilent 83630L T est Record

T able 2-35. T est Record for 83630L (4 of 5)

No. T est Description Sp ec. Results Sp ec. Uncertain t y
0.01 to 26.5 GHz +8.0 dBm 1.7 dB
0.01 to 20 GHz +11.5 dBm 1.7 dB
100 MHz Fixed 60 dBc 1 dB
100 MHz O set 60 dBc 1 dB
LO F eedthrough 60 dBc 1 dB
Agilent 8360 P erformance T ests
Agilent 83630L T est Record

T able 2-35. T est Record for 83630L (5 of 5)

No. T est Description Sp ec. Results Sp ec. Uncertaint y
120 Hz Spur 55 dBc 1 dB
180 Hz Spur 55 dBc 1 dB
240 Hz Spur 55 dBc 1 dB
120 Hz Spur 55 dBc 1 dB
180 Hz Spur 55 dBc 1 dB
240 Hz Spur 55 dBc 1 dB
100 Hz to < 1 kHz 70 dBc 2 dB
1 kHz to < 10 kHz 78 dBc 2 dB
10 kHz to < 100 kHz 86 dBc 2 dB
100 kHz 107 dBc 2 dB
100 Hz to < 1 kHz 70 dBc 2 dB
1 kHz to < 10 kHz 78 dBc 2 dB
10 kHz to < 100 kHz 86 dBc 2 dB
100 kHz 107 dBc 2 dB
100 Hz to < 1 kHz 60 dBc 2 dB
1 kHz to < 10 kHz 68 dBc 2 dB
10 kHz to < 100 kHz 76 dBc 2 dB
100 kHz 97 dBc 2 dB
1 3 dB if the Invertron is not used.
2-158 P erformance T ests
Agilent 83630L T est Record

T able 2-36. T est Record for Agilent 83640L (1 of 5)
Mo del Am bien t temp erature
Serial Num b er Relative humidit y %
Options Line frequency Hz (nominal)
Firm w are Revision
Sp ecial Notes:
Agilent 8360 P erformance T ests
Agilent 83640L T est Record

T able 2-36. T est Record for 83640L (2 of 5)
Mo del Rep ort Num b er Date
T est Equipmen t Used Mo del Numb er T race Numb er Cal Due Date
2-160 P erformance T ests
Agilent 83640L T est Record

T able 2-36. T est Record for 83640L (3 of 5)

No. T est Description Sp ec. Results Sp ec. Uncertain t y
< 2 GHz 0.6 dB 0.6 dB 0.22 dB
2 and 20 GHz 0.7 dB 0.7 dB 0.35 dB
> 20 and 40 GHz 0.9 dB 0.9 dB 0.38 dB
2 and 20 GHz 1.2 dB 0.25 dB
> 20 and 40 GHz 1.8 dB 0.27 dB
0.01 to 40 GHz +6.0 dBm 1.4 dB
0.01 to < 26.5 GHz +10.0 dBm 1.8 dB
0.01 to < 20 GHz +10.0 dBm 1.8 dB
Agilent 8360 P erformance T ests
Agilent 83640L T est Record

T able 2-36. T est Record for the 83640L (4 of 5)

No. T est Description Sp ec. Results Sp ec. Uncertaint y
0.01 to 40 GHz +4 dBm 1.4 dB
0.01 to < 26.5 GHz +8 dBm 1.8 dB
0.01 to < 20 GHz +8.5 dBm 1.8 dB
100 MHz Fixed 60 dBc 1 dB
100 MHz O set 60 dBc 1 dB
LO F eedthrough 60 dBc 1 dB
2-162 P erformance T ests
Agilent 83640L T est Record

T able 2-36. T est Record for the 83640L (5 of 5)

No. T est Description Sp ec. Results Sp ec. Uncertain t y
100 Hz to < 1 kHz 70 dBc 2 dB
1 kHz to < 10 kHz 78 dBc 2 dB
10 kHz to < 100 kHz 86 dBc 2 dB
100 kHz 107 dBc 2 dB
100 Hz to < 1 kHz 70 dBc 2 dB
1 kHz to < 10 kHz 78 dBc 2 dB
10 kHz to < 100 kHz 86 dBc 2 dB
100 kHz 107 dBc 2 dB
100 Hz to < 1 kHz 60 dBc 2 dB
1 kHz to < 10 kHz 68 dBc 2 dB
10 kHz to < 100 kHz 76 dBc 2 dB
100 kHz 97 dBc 2 dB
1 3 dB if the Invertron is not used.
Agilent 8360 P erformance T ests
Agilent 83640L T est Record

T able 2-37. T est Record for Agilent 83650L (1 of 6)
Mo del Am bien t temp erature
Serial Num b er Relative humidit y %
Options Line frequency Hz (nominal)
Firm w are Revision
Sp ecial Notes:
2-164 P erformance T ests
Agilnet 83650L T est Record

T able 2-37. T est Record for 83650L (2 of 6)
Mo del Rep ort Num b er Date
T est Equipmen t Used Mo del Numb er T race Numb er Cal Due Date
Agilent 8360 P erformance T ests
Agilnet 83650L T est Record

T able 2-37. T est Record for 83650L (3 of 6)

No. T est Description Sp ec. Results Sp ec. Uncertaint y
< 2 GHz 0.6 dB 0.6 dB 0.22 dB
2 and 20 GHz 0.7 dB 0.7 dB 0.35 dB
> 20 and 40 GHz 0.9 dB 0.9 dB 0.38 dB
> 40 GHz 1.7 dB 1.7 dB 0.79 dB
2 and 20 GHz 1.2 dB 0.25 dB
> 20 and 40 GHz 1.6 dB 0.38 dB
2-166 P erformance T ests
Agilnet 83650L T est Record

T able 2-37. T est Record for 83650L (4 of 6)

No. T est Description Sp ec. Results Sp ec. Uncertain t y
0.01 to 50 GHz +2.5 dBm 2.3 dB
0.01 to < 40 GHz +5.0 dBm 1.3 dB
0.01 to < 26.5 GHz +10.0 dBm 1.7 dB
0.01 to < 20 GHz +10.0 dBm 1.7 dB
0.01 to 50 GHz 0 dBm 2.3 dB
0.01 to < 40 GHz +3.0 dBm 1.3 dB
0.01 to < 26.5 GHz +8.0 dBm 1.7 dB
0.01 to < 20 GHz +8.5 dBm 1.7 dB
Agilent 8360 P erformance T ests
Agilnet 83650L T est Record

T able 2-37. T est Record for 83650L (5 of 6)

No. T est Description Sp ec. Results Sp ec. Uncertaint y
100 MHz Fixed 60 dBc 1 dB
100 MHz O set 60 dBc 1 dB
LO F eedthrough 60 dBc 1 dB
120 Hz Spur 55 dBc 1 dB
180 Hz Spur 55 dBc 1 dB
240 Hz Spur 55 dBc 1 dB
120 Hz Spur 55 dBc 1 dB
180 Hz Spur 55 dBc 1 dB
240 Hz Spur 55 dBc 1 dB
1 3 dB if the Invertron is not used.
2-168 P erformance T ests
Agilnet 83650L T est Record

T able 2-37. T est Record for 83650L (6 of 6)

No. T est Description Sp ec. Results Sp ec. Uncertaint y
100 Hz to < 1 kHz 70 dBc 2 dB
1 kHz to < 10 kHz 78 dBc 2 dB
10 kHz to < 100 kHz 86 dBc 2 dB
100 kHz 107 dBc 2 dB
100 Hz to < 1 kHz 70 dBc 2 dB
1 kHz to < 10 kHz 78 dBc 2 dB
10 kHz to < 100 kHz 86 dBc 2 dB
100 kHz 107 dBc 2 dB
100 Hz to < 1 kHz 60 dBc 2 dB
1 kHz to < 10 kHz 68 dBc 2 dB
10 kHz to < 100 kHz 76 dBc 2 dB
100 kHz 97 dBc 2 dB
Agilent 8360 P erformance T ests
Agilnet 83650L T est Record



How T o Use This Chapter
P erform these adjustmen ts only if directed b y the A gilent T e chnolo gies 8360 B-Series
Swept Signal Gener ator/ 8360 L-Series Swept CW Gener ator T r oublesho oting Guide or if a
p erformance test fails.
F old-out men u maps are lo cated in Chapter 6.
Some men us ha v e more than one page of softkeys. Select the
More softkey to view the next
page of softkeys.
More is not included in the k eystrok es given in these pro cedures.
In all cases where y ou are instructed to preset the syn thesizer, use the factory
preset mo de only .
If a password is set, it will ha v e to b e disabled b efore the adjustmen t men u
can b e accessed. See Chapter 4 for instructions.
The equipmen t required to p erform the adjustmen ts in this c hapter is listed in Chapter 1.
Y ou may use an y equipmen t that satis es the critical sp eci cations given. Use a non-metal lic
adjustmen t to ol.
This instrumen t has b een designed in accordance with in ternational safety standards.
Ho w ev er, for safe op eration follow the cautions and w arnings in this man ual.
Agilent 8360 Adjustments 3-1
Adjustments · Menu Maps · Equipment Required

These servicing instructions are for use b y qualified personnel only . T o a v oid
electrical shock, do not perform any servicing unless y ou are qualified to do
so.
The opening of co v ers or remo v al of parts is lik ely to expose dangerous
v oltages. Disconnect the instrument from all v oltage sources while it is being
opened.
The detachable po w er cord is the instrument disconnecting device. It
disconnects the mains circuits from the mains supply before other parts of
the instrument. The front panel switch is only a standb y switch and is not a
LINE switch.
The po w er cord is connected to internal capacitors that may remain liv e for 5
seconds after disconnecting the plug from its po w er supply .
This is a Safety Class I product (pro vided with a protectiv e earthing ground
incorporated in the po w er cord). The mains plug shall only be inserted in a
sock et outlet pro vided with a protectiv e earth contact. Any interruption of the
protectiv e conductor, inside or outside the instrument, is lik ely to mak e the
instrument dangerous. Intentional interruption is prohibited.
For continued protection against fire hazard replace line fuse only with same
type and rating (F 5A/250 V). The use of other fuses or material is prohibited.
Always use the three-prong ac p ow er cord supplied with this instrumen t.
F ailure to ensure adequate earth grounding b y not using this cord may cause
instrumen t damage.
Before switching on this pro duct, make sure that the line v oltage selector
switch is set to the v oltage of the p ow er supply and the correct fuse is
installed. Assure the supply v oltage is in the sp eci ed range.
3-2 Adjustments Agilent 8360

The follo wing is a list of this c hapter's adjustmen ts and the page n um b er that eac h
adjustmen t is lo cated on:
(this adjustmen t may b e replaced with the automated
p ow er atness adjustmen t in Chapter 5.)
Adjustments for 8360 B-Series Only
The adjustmen ts that apply to the
8360 B-series swept signal generators
only are marked with the following symb ol: y
F or the following automated adjustmen ts, refer to Chapter 5, \Automated T ests":
YO Delay Adjustmen t
P ow er Flatness
Agilent 8360 Adjustments 3-3
List of Adjustments · ADC Adjustmen t

This pro cedure adjusts the frequency accuracy of the in ternal 10 MHz time base. This
adjustmen t should b e done on a regular basis if absolute frequency accuracy is imp ortant (see
Figure 3-3.)
F or b est accuracy , readjust the 10 MHz timebase oscillator after the syn thesizer has b een
on or in standb y for 24 hours. See \Accuracy V ersus Adjustmen t In terv al," following this
adjustmen t, for information on ho w to determine a p erio dic adjustmen t sc hedule.
After the timebase is adjusted, the timebase frequency should stay within the aging rate if the
follo wing things happ en:
The time base ov en do es not co ol do wn.
The instrumen t k eeps the same orientation with resp ect to the earth's magnetic eld.
The instrumen t stays at the same altitude.
The instrumen t do es not receiv e an y mec hanical sho c k.
If the time base ov en co ols (the instrumen t is disconnected from A C p ow er), y ou may ha v e to
readjust the time base frequency after a new w arm-up cycle. T ypically , ho w ev er, the time base
frequency returns to within 1 Hz of the original frequency .
Y ou can adjust the in ternal timebase after reconnecting A C p ow er for 10
minutes, but for b est accuracy , test again after the instrumen t has b een on or
in standb y for 24 hours.
F requency c hanges, due either to a c hange in orientation with resp ect to the earth's magnetic
eld or to a c hange in altitude, are usually eliminated when the instrumen t is returned to its
original p osition. A frequency c hange due to mec hanical sho c k usually app ears as a xed
frequency error.
Description and Procedure · 3-4 Adjustments · 10 MHz Standard

up for one hour.

Figure 3-1. 10 MHz Standard Adjustment Setup
Digitizing Oscilloscop e HP/Agilent 54111D
F requency Standard HP/Agilent 5061A
If the oscilloscop e do es not ha v e a 50 input imp edance, connect c hannel 1
through a 50 feedthrough.
Channel 1:
Channel 2:
Timebase:
Trigger:
Trigger Source Trig 3
Display:
Display Mode Real Time
syn thesizer's in ternal standard. The w av eform will app ear to drift.

horizontal mov emen t of the oscilloscop e w av eform.

Figure 3-2. 10 MHz Standard Adjustment Location
3-6 Adjustments · 10 MHz Standard

Related P erformance T ests
In Case of Difficulty
automatically c ho oses the external standard as the reference if one is connected to the 10
MHz REF INPUT.)
Swept CW Gener ator T r oublesho oting Guide .
Agilent 8360 Adjustments · 10 MHz Standard

Accuracy V ersus Adjustment Interv al
Figure 3-3 sho ws the required adjustmen t in terv al to maintain a given accuracy . If y ou kno w
the aging rate, y ou can determine a more precise adjustmen t in terv al.

Figure 3-3. Accuracy V ersus Adjustment Interv al
(T o determine the aging rate, see \Performance T ests.")
calibration in terv al measured in da ys.
Y ou can also determine the minimum calibration in terv al using the following formula:
maximum allo w able error
= calibration in terv al in da ys
kno wn aging rate (p er da y)
An Alternate Method · 3-8 Adjustments · 10 MHz Standard

No test equipmen t is required for this adjustmen t.
This is an automatic adjustmen t. The syn thesizer's in ternal DVM measures the sweep
ramp v oltage at the end of the sweep. The syn thesizer then sets the sweep time calibration
constants to calibrate the sweep ramp.
. Let the syn thesizer w arm up for at least one
hour.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SERVICE
The syn thesizer will measure, calculate, and store the sweep time calibration constants.
When the adjustmen t is complete, the follo wing message app ears on the display:
Sweep Ramp Cal Completed
Calibration constants were modified.
store the calibration constants as protected data
(in EEPR OM).
Related P erformance T ests
Sw ept F requency Accuracy
In Case of Difficulty
Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series
Swept CW Gener ator T r oublesho oting Guide .

No test equipmen t is required for this pro cedure.
There are three sampler assembly adjustmen ts:
In the sampler match adjustmen t, as the in ternal DVM monitors the lo op input to the
sampler, trim capacitors are adjusted to optimize the sampler match ov er the full frequency
range of the lo op.
The 200 MHz lo op gain adjustmen t is automatic. When this adjustmen t is selected in the
A6 adjustmen t men u, the syn thesizer selects in ternal calibration constants for optimum lo op
gain ov er the full frequency range of the lo op.
In the IF gain adjustmen t, the IF ampli er gain is adjusted as the in ternal DVM monitors
the IF output lev el to the YO lo op.
b et w een FL2 and the A6 assembly .

Figure 3-4. FL2 Location
on the analog extender b oard.
. Let it w arm up for at least one hour.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SERVICE

200 MHz Loop V CO Tune Check
In the A6 adjustmen t men u, select the A6 V CO tune c hec k
A6 VCO Tune .
Ctrim and C60 are OK.
If this message is not display ed, refer to troublesho oting.
If the three tuning v oltages at the b ottom of the display (V200, V avg, and
V220) are all equal, the 200 MHz phase-lo c k ed lo op is unlo c k ed.
Done .
In the A6 adjustmen t men u, select the A6 sampler match adjustmen t,
A6 Smplr Match .
The sampler match display is a graphic represen tation of the sampler match
ov er the full frequency range of the 200 MHz lo op. The more asterisks
display ed at a frequency , the more output signal (and the b etter the match) at
that frequency . Make the adjustmen t for b est match ov er the full frequency
range of the lo op.

Figure 3-5. Sampler Match Adjustment Locations
Done .

200 MHz Loop Gain Adjustment
In the A6 adjustmen t men u, select the A6 lo op gain adjustmen t,
A6 Loop Gain . The
syn thesizer automatically p erforms the adjustmen t. When the message Calibration
Constants should be SAVED is display ed, the adjustmen t is complete. Con tinue with the
next step.
all cables.
T urn the syn thesizer on. Select
A6 IF Gain .

Figure 3-6. IF Gain Adjustment Location
Done .
to store the calibration constants as protected data (in EEPR OM).
Related P erformance T ests
In Case of Difficulty
analog extender b oard.
coax cables (including semi-rigid) are connected and that no unlo c k ed message is display ed.

order of adjustmen ts.
Swept CW Gener ator T r oublesho oting Guide .
Agilent 8360 Adjustments · Sampler Assembly

No test equipmen t is required for this pro cedure.
As the in ternal DVM monitors the +10 V reference, y ou make the adjustmen t.
. Let the instrumen t w arm up for at least one
hour.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SERVICE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Select the A13 +10 V reference adjustmen t,
A13+10V Ref .

Figure 3-7. YO Driv er +10 V Reference Adjustment Location
Done .
Related P erformance T ests
In Case of Difficulty
Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series
Swept CW Gener ator T r oublesho oting Guide .
YO Driv er +10 V Reference

No test equipmen t is required for these adjustmen ts.
F or the gain adjustmen t, the syn thesizer's in ternal DVM monitors the YO lo op error v oltage
at tw o frequencies (p oints B and C in Figure 3-8). When y ou select the gain p oten tiometer
adjustmen t, the syn thesizer displays the di erence in v oltage b et w een p oints B and C as a
\needle" with arrows on either side. By adjusting A13R54 to cen ter the \needle", y ou equalize
the v oltage b et w een the tw o frequency p oints.
The linearit y adjustmen t has tw o parts (eac h of whic h is selected and adjusted similarly to the
gain adjustmen t). The low break p oint adjustmen t equalizes the v oltage b et w een p oints A and
B in Figure 3-8. The high break p oint adjustmen t equalizes the v oltage b et w een p oints C and
D.
Y ou m ust adjust the gain rst; then adjust the break p oints.

Figure 3-8. YO Gain and Linearity Break P oints
. Let it w arm up for at least one hour.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SERVICE
If the follo wing message is display ed: Adjust A13 pot until "locked".,
adjust the p oten tiometer un til the lo c k ed indication is display ed. Con tin ue
with step 3.
Gain Pot .
YO Driv er Gain and Linearity


Figure 3-9. YO Gain and Linearity Adjustment Locations
Select the low break p oint p oten tiometer adjustmen t,
Lo Bk Pot .
Select the high break p oint p oten tiometer adjustmen t,
Hi Bk Pot .
Done .
to store the calibration constants as protected data (in EEPR OM).
Related P erformance T ests
In Case of Difficulty
Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series
Swept CW Gener ator T r oublesho oting Guide .
YO Driv er Gain and Linearity
Select · Self-tests · 3-16 Adjustments

No test equipmen t is required for this pro cedure.
. Let the instrumen t w arm up for at least one
hour.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SERVICE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Select the A5 lo op gain adjustmen t,
A5 Loop Gain .

Figure 3-10. Switch and Adjustment Locations
Done .
Set all A5 switches to op en and select
Done .
Related P erformance T ests
In Case of Difficulty
when the adjustmen t is complete.
Swept CW Gener ator T r oublesho oting Guide .

This pro cedure maximizes RF p ow er b y tracking the tuned lter in the ampli er/multiplier
to the RF output frequency . The ampli er/ lter output is also monitored. Initial tracking
is done in single band sweeps at slow sweep sp eeds to eliminate the e ects of bandcross and
h ysteresis in the tuning coil. (Those will b e corrected with dela y and risetime calibration
constants.) A squegging clamp adjustmen t limits the p ow er in to the ampli er/m ultiplier
and is adjusted in the ampli er/ lter adjustmen ts for maximum output p ow er without
squegging. A \squegg" is a distortion in the RF output caused b y to o m uc h p ow er to the
ampli er/multiplier and c haracterized b y a p ow er drop out in a p ortion of the trace, or
a p ow er drop out ov er a broad frequency range. The SRD bias adjustmen ts are made to
optimize the eciency of the frequency m ultiplication. The YTM dela y comp ensation
adjustmen ts maximize p ow er for fast single- and m ulti-band sweeps.
If y ou ha v e replaced either the YO, the YO driver, or the m ultiplier/ lter
driver, rst initiate an auto tracking (press
USER CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Auto Track .) If auto tracking passes and the instrumen t passes the
\Maximum Lev eled P ow er" p erformance test, do not con tinue with this
pro cedure. If auto tracking fails, (an error message is display ed), con tinue with
this pro cedure.

The follo wing explanations are provided for a b etter understanding of the ampli er/multiplier
adjustmen ts. Refer to these explanations as often as necessary . They will help make these
adjustmen ts easier. (The adjustmen t pro cedure follo ws.)
Offset and Gain Adjustments.
F or all o set and gain adjustmen ts, adjust thr ough the bandpass. Keep adjusting un til the
p ow er p eaks and then drops o . Then reset the adjustmen t to the p eak ed p oint.
F or all o set and gain adjustmen ts, if p ow er stays p eak ed ov er sev eral calibration constant
v alues, set the calibration constant to the middle v alue of the p eak ed range.
Adjusting to the Center of the P assband.
The ampli er/multiplier has a tracking lter on its output. Its purp ose is to pass the
selected RF output frequency and reject harmonics. The ob jective of this pro cedure is to
track the bandpass lter with the RF output frequency ov er a v ariety of sweep conditions
and frequency ranges. E ects of the auxiliary output may mask the true cen ter of the
bandpass. Always tune to the low er p eak (lo w er calibration constant v alue).
SRD Bias Adjustment.
Adjust these calibration constants into sque gging and then bac k o un til p eak p ow er is
ac hiev ed.
After reac hing p eak p ow er, decrease the calibration constant v alue b y 15 coun ts when
adjusting the A term and decrease p ow er b y 1/2 dB (approximately 1/3 division) when
adjusting the B term.
Single Band Delay Compensation Adjustments.
If a drop in p ow er greater than 1 division o ccurs when in single sweep, increase the A
dela y term to put the A term at the high end of the p eak. F or example, for the following
calibration constant v alues:
Start of the p eak = 1800
Middle of the p eak = 1950
End of the p eak = 2100
Set the A term to 2050
Setting the F astest Sw eep Time.
Setting the sweep time to 0 milliseconds sets the syn thesizer to the fastest sweep time for
the frequency range b eing swept. Eac h time the frequency range is c hanged, the sweep time
m ust b e reset to 0 ms in order to maintain the fastest sweep time.
Sw eep Speed Related Adjustments.
Delay comp ensation adjustmen ts are a ected b y sweep sp eed. The adjustmen ts are
p erformed at fast sweep sp eeds whic h are the w orst case.
All other adjustmen ts are p erformed at slow sweep sp eeds; they are not a ected b y sweep
sp eed.
If y ou ha v e a p ow er problem that o ccurs only for fast sweep rates, it is probably a ected b y
dela y comp ensation or risetime adjustmen ts.
Adjustment Help · Agilent 8360 Adjustments · Amplifier/Multiplier Adjustments

YTM Bandcross Delay T erms.
The YTM Bx Dly terms are either o set or gain terms used only during m ulti-band sweeps.
A1 denotes the o set term for band 1. B1 denotes the gain term for band 1. The o set
(A) should b e adjusted to maximize p ow er at the b eginning of the band. The gain (B) is
adjusted to maximize p ow er tow ard the end of the band.
General T echniques.
In some adjustmen ts p ow er may p eak twice. This is caused b y the auxiliary output masking
the true p eak. Always adjust to the rst p eak (low est v alue of the calibration constant).
If the adjustmen t has a wide range without an y apparen t improv emen t. Lea v e it set to the
default v alue.
3-20 Adjustments · Amplifier/Multiplier Adjustments

In this pro cedure, some calibration constants and p oten tiometers may b e
set to default conditions and left. This is normal. These b oards con tain
adjustmen ts for di eren t vin tages of the 8360 family . Not all adjustmen ts are
used eac h time.
v ersus B sweep capabilit y . P ow er on all the instrumen ts and let them w arm up for at least
one hour.
.

Figure 3-11. Amplifier/Multiplier Adjustments Setup
Analog Oscilloscop e HP/Agilent 1740A
Oscilloscop e Prob es HP/Agilent 10431A
Detector HP/Agilent 33330D (Agilent 83620B /22B/24B,
Agilent 83623B/L, and Agilent 83630B/L)
Detector HP/Agilent 33330E (Agilent 83640B/L & Agilent 83650B/L)
A tten uator 10 dB HP/Agilent 8493C Option 010 (83620B /22B/24B,
A tten uator 10 dB HP/Agilent 8490D Option 010
constants. On the syn thesizer, press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SERVICE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Select Cal En ter the n um b er of the rst calibration constant from T able 3-1 and
terminate the en try with
. When using the front panel em ulator softw are, pressing
the arrow k eys will allo w y ou to v ary the calibration constants. The calibration constant
sho wn at the top of the \display" window is the active constant.
Modify Cal En ter the default v alue from T able 3-1 and terminate the en try with
.
Using the up/do wn arrow k eys, select eac h of the rest of the calibration constants in
T able 3-1 and set their default v alues b y en tering the v alue on the n umeric k eypad.
T erminate y our en try with the
k ey .
T able 3-1. Calibration Constant Descriptions
Numb er Description Default
105 SRD Bias A; Band 1 255 Maximize p ow er over low er 30%
106 SRD Bias A; Band 2 130
of band. Sw eep sp eed is not a
factor. Adjust for 0.5 dB
107 SRD Bias A; Band 3 130
b elow maxim um p ow er.
108 SRD Bias A; Band 4 130
109 SRD Bias A; Band 5 130
110 SRD Bias A; Band 6 120
111 SRD Bias A; Band 7 120
118 SRD Bias B; Band 1 255 Maximize p ow er over higher 30%
119 SRD Bias B; Band 2 150
of band. Sw eep sp eed is not a
factor. Adjust for 0.5 dB b elow
120 SRD Bias B; Band 3 150
maxim um p ow er.
121 SRD Bias B; Band 4 150
122 SRD Bias B; Band 5 160
123 SRD Bias B; Band 6 160
124 SRD Bias B; Band 7 160
131 Squegg Clamp 1A 255 Default v alues are not altered.
132 Squegg Clamp 2A 255
Num b ers should remain
unchanged (#131 is adjusted
133 Squegg Clamp 3A 255
in another pro cedure).
134 Squegg Clamp 4A 255
135 Squegg Clamp 5A 255
136 Squegg Clamp 6A 255
137 Squegg Clamp 7A 255
144 YTM Delay T erm A Hrm 1 1000 Default v alues are not altered.
Num b ers should remain
145 YTM Delay T erm A Hrm 2 1000
unchanged. Maximize p ow er over
146 YTM Delay T erm A Hrm 3 1000
rst 20% of band. A ects
fast sw eeps only . Single sw eep
147 YTM Delay T erm A Hrm 4 1000
mo de is also critical.
148 YTM Delay T erm A Hrm 5 1000
149 YTM Delay T erm A Hrm 6 1000
150 YTM Delay T erm A Hrm 7 1000

T able 3-1. Calibration Constant Descriptions (continued)
Numb er Description Default
157 YTM Delay T erm B Hrm 1 300 Default v alues are not altered.
Num b ers should remain
158 YTM Delay T erm B Hrm 2 300
unchanged. Maximize p ow er
159 YTM Delay T erm B Hrm 3 300
over higher 30% of band.
160 YTM Delay T erm B Hrm 4 300
A ects fast sw eeps only .
161 YTM Delay T erm B Hrm 5 300
162 YTM Delay T erm B Hrm 6 300
163 YTM Delay T erm B Hrm 7 300
170 YTM Bx Dly T erm A1 1000 Default v alues for all
terms except A3 and A4.
171 YTM Bx Dly T erm A2 1000
Delay comp ensation for
172 YTM Bx Dly T erm A3 1000
m ulti-band sw eeps only .
173 YTM Bx Dly T erm A4 1000
Adjust for maxim um p ow er at
low er 50% of band. A ects
174 YTM Bx Dly T erm A5 1000
fast sw eep sp eeds only .
175 YTM Bx Dly T erm A6 1000
176 YTM Bx Dly T erm A7 1000
183 YTM Bx Dly T erm B1 300 Default v alues for all
terms except B3 and B4.
184 YTM Bx Dly T erm B2 300
Delay comp ensation for
185 YTM Bx Dly T erm B3 300
m ulti-band sw eeps only .
186 YTM Bx Dly T erm B4 300
Adjust for maxim um p ow er at
low er 50% of band. A ects
187 YTM Bx Dly T erm B5 300
fast sw eep sp eeds only .
188 YTM Bx Dly T erm B6 300
189 YTM Bx Dly T erm B7 300
195 YTM Kic k Threshold 110 Default v alues are not altered.
Num b ers should remain
196 YTM CW Kic k Max 600
unchanged.
197 YTM Mono Band Kic k 2000
198 YTM Stereo Band Kic k 2000
199 YTM Slew Rate 400
200 YTM Slew Max 50
201 YTM Slew Min 0
202 YTM Neg Kic k W ait 2
203 YTM Fwd Kic k Pct 50
204 YTM Fwd Kic k W ait 30

T able 3-1. Calibration Constant Descriptions (continued)
Numb er Description Default
205 YTM Rise; Band 1 15 Minimize p ow er drop outs at the
start of band. Adjustmen t is only
206 YTM Rise; Band 2 A1 20
e ectiv e at fast sw eep sp eeds.
207 YTM Rise; Band 2 B1 2
Adjusted in the Ampli er/
208 YTM Rise; Band 2 A2 15
Filter Adjustments.
209 YTM Rise; Band 3 A1 70
210 YTM Rise; Band 3 B1 7
211 YTM Rise; Band 3 A2 15
212 YTM Rise; Band 3 B2 15
213 YTM Rise; Band 3 A3 15
214 YTM Rise; Band 4 15
215 YTM Rise; Band 5 15
216 YTM Rise; Band 6 10
217 YTM Rise; Band 7 10
225 *YTM Gain Band 1 2048 Maximize p ow er over higher 10%
226 *YTM Gain Band 2 2048
of band. Adjust at slow sw eep
sp eeds.
227 *YTM Gain Band 3 2048
228 *YTM Gain Band 4 2048
229 *YTM Gain Band 5 2048
230 *YTM Gain Band 6 2048
231 *YTM Gain Band 7 2048
238 *YTM O set Band 1 2048 Maximize p ow er over low er 10%
of band. Adjust at slow sw eep
239 *YTM O set Band 2 2048
sp eeds.
240 *YTM O set Band 3 2048
241 *YTM O set Band 4 2048
242 *YTM O set Band 5 2048
243 *YTM O set Band 6 2048
244 *YTM O set Band 7 2048
248 YTM B2 O set O set 0 Default v alues are not altered.
Num b ers should remain
unchanged.

T able 3-1. Calibration Constant Descriptions (continued)
Numb er Description Default
373 YTM Dly T erm C Hrm 1 0 Default v alues are not altered.
Num b ers should remain
374 YTM Dly T erm C Hrm 2 0
unchanged.
375 YTM Dly T erm C Hrm 3 0
376 YTM Dly T erm C Hrm 4 0
377 YTM Dly T erm C Hrm 5 0
378 YTM Dly T erm C Hrm 6 0
379 YTM Dly T erm C Hrm 7 0
382 YTM Bx Dly T erm C1 0 Default v alues are not altered.
Num b ers should remain
383 YTM Bx Dly T erm C2 0
unchanged.
384 YTM Bx Dly T erm C3 0
385 YTM Bx Dly T erm C4 0
386 YTM Bx Dly T erm C5 0
387 YTM Bx Dly T erm C6 0
525 Hib ernation Time 90 Default v alues are not altered.
526 Bandcross 1 42
Num b ers should remain
unchanged.
527 Bandcross 2 32
528 Bandcross 3 28
529 Bandcross 4 26
530 Bandcross 5 28
531 Bandcross 6 20
532 Bandcross 7 20
Band 1 Adjustment.
2.0
SWEEP TIME
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Leveling Point Ext Det (asterisk on)
POWER LEVEL
Channel A:

Adjust Channel B o set, v olts/division, and horizontal p osition con trols for a trace that
lls the full horizontal display . Adjust the v ertical p osition with the A v ertical p osition
con trol.
Then adjust A12R69, YTMG, to maximize high end p ow er (last tw o horizontal display
divisions).
Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace
shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 6 through 8 again.
Change calibration constant #238 bac k to its original v alue.
Band 2 Adjustment.
STOP
13.5
high end p ow er may p eak twice. If it do es, adjust the calibration constant to the low er
p eak (where the v alue of the calibration constant is the low er n um b er).
P ow er may p eak twice. Adjust to the low er p eak.
a. Incremen t calibration constant #106, SRD Bias A; Band 2, to maximize low end
p ow er. Then decrease the calibration constant v alue b y 15 coun ts.
b. Incremen t calibration constant #119, SRD Bias B; Band 2, to maximize high end
p ow er. Then decrease the p ow er b y one third division ( 1/2 dB).
Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace
shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 12 through 15.
Change calibration constant #239 bac k to its original v alue.
Band 3 Adjustment.
13.5
There are tw o sets of p oten tiometers on the A12 assembly with iden tical
names. Make c ertain y ou are adjusting the correct ones.


Figure 3-12. A12 Assembly P otentiometer Locations
P ow er may p eak twice. Adjust to the low er p eak.
end p ow er may p eak twice. If it do es, adjust the p oten tiometer to the low er p eak (where
the adjustmen t is furthest coun terclo c kwise).
a. Incremen t calibration constant #107, SRD Bias A; Band 3, to maximize low end
p ow er. Then decrease the calibration constant v alue b y 15 coun ts.
b. Incremen t calibration constant #120, SRD Bias B; Band 3, to maximize high end
p ow er. Then decrease the p ow er b y one third division ( 1/2 dB).
Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace
shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 18 through 23.
Change calibration constant #240 bac k to its original v alue.
Band 4 Adjustment (83630B/L, 83640B/L, and 83650B/L only).
26.5
( 83630B /L and 83650B /L )
STOP
25.5
A12R23 B4BP1 fully clo c kwise.
A12R24 BP2 fully clo c kwise.
A12R26 B4S1 fully clo c kwise.
A12R27 S2 fully clo c kwise.

low end. P ow er may p eak twice. Adjust to the low er p eak.
division on the display . P ow er may p eak twice. Adjust to the low er p eak.
If there is no noticeable increase in p ow er in the next tw o adjustmen ts, leav e
the p oten tiometers at their preset p osition (fully clo c kwise).
low er p eak.
Stop when the p ow er b egins to decrease. P ow er may p eak twice. Adjust to the low er
p eak.
a. Incremen t calibration constant #108, SRD Bias A; Band 4, to maximize low end
p ow er. Then decrease the calibration constant v alue b y 15 coun ts.
b. Incremen t calibration constant #121, SRD Bias B; Band 4, to maximize high end
p ow er. Then decrease the p ow er b y one third division ( 1/2 dB).
Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace
shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 26 through 34.
Change calibration constant #241 bac k to its original v alue.
Band 5 Adjustment (83640B/L and 83650B/L only).
25.5
ST ART
26.5
ST ART
STOP
( 83640B /L)
STOP
33.5
end of band 5. The high end p ow er may p eak twice. If it do es, adjust the calibration
constant to the low er p eak (where the v alue of the calibration constant is the low er
n um b er).
P ow er may p eak twice. Adjust to the low er p eak.
a. Incremen t calibration constant #109, SRD Bias A; Band 5, to maximize low end
p ow er. Then decrease the calibration constant v alue b y 15 coun ts.
b. Incremen t calibration constant #122, SRD Bias B; Band 5, to maximize high end
p ow er. Then decrease the p ow er b y one third division ( 1/2 dB).
Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace
shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 37 through 40.
Change calibration constant #242 bac k to its original v alue.

Band 6 Adjustment (83640B/L and 83650B/L only).
ST ART
33.5
ST ART
STOP
( 83640B /L)
STOP
( 83650B /L)
end of band 6. The high end p ow er may p eak twice. If it do es, adjust the calibration
constant to the low er p eak (where the v alue of the calibration constant is the low er
n um b er).
P ow er may p eak twice. Adjust to the low er p eak.
a. Incremen t calibration constant #110, SRD Bias A; Band 6, to maximize low end
p ow er. Then decrease the calibration constant v alue b y 15 coun ts.
Incremen t calibration constant #123, SRD Bias B; Band 6, to maximize high end
p ow er. Then decrease the p ow er b y one third division ( 1/2 dB).
Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace
shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 43 through 46.
Change calibration constant #243 bac k to its original v alue.
Band 7 Adjustment (83650B/L only).
STOP
end of band 7. The high end p ow er may p eak twice. If it do es, adjust the calibration
constant to the low er p eak (where the v alue of the calibration constant is the low er
n um b er).
P ow er may p eak twice. Adjust to the low er p eak.
a. Incremen t calibration constant #111, SRD Bias A; Band 7, to maximize low end
p ow er. Then decrease the calibration constant v alue b y 15 coun ts.
b. Incremen t calibration constant #124, SRD Bias B; Band 7, to maximize high end
p ow er. Then decrease the p ow er b y one third division ( 1/2 dB).
Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace
shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 49 through 52.
Change calibration constant #244 bac k to its original v alue.
Multi-band Amplifier/Multiplier Delay Compensation.

P ow er drop outs at the start of eac h band are typically remov ed with the YTM
Risetime calibration constants. These will b e adjusted later.
2.0
STOP
SWEEP TIME
The syn thesizer in tensity marker is set to iden tify the bandcross frequency
(13.5 GHz). Reduce oscilloscop e trace in tensity to iden tify the marker. Y ou
can also iden tify bandcrosses b y temp orarily removing the cable from the
Z-AXIS BLANK/MKRS connector on the syn thesizer rear panel. Make sure
this cable is connected when making adjustmen ts.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
13.5
(start of band 3)
end of band 3 (adjust the low est amplitude p ortion of the band).
end of band 3.
p ow er c hanges less than 2/3 division ( 1 dB). If the c hange in p ow er is greater than 2/3
division, readjust the A term of the dela y adjustmen t #172 for low end problems, or the B
term of the dela y adjustmen t #185 for high end problems.
SWEEP TIME
sev eral times to initiate sev eral sweeps. If a drop in p ow er greater than 1
division o ccurs when in single sweep, increase the A term of the dela y adjustmen t #172 to
put the A term at the high end of the p eak. (See \Adjustmen t Help.")
Band 4 Multi-band Amplifier/Multiplier Delay Compensation (83630B/L, 83640B/L, and 83650B/L
only).
2.0
26.5
( 83630B /L and 83650B /L )
STOP
25.5
CONT
SWEEP TIME
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
(start of band 4)
of band 4.
end of band 4.
p ow er c hanges less than 2/3 division ( 1 dB) across b oth bands. If the c hange is greater
than 2/3 division, readjust the appropriate delay term.

sev eral times to initiate sev eral sweeps. If a drop in p ow er greater than 1
division o ccurs when in single sweep, readjust the appropriate dela y term for the lo cation
of the problem.
CONT
SWEEP TIME
sev eral times to initiate sev eral sweeps. If a drop in p ow er greater than 1/3
division o ccurs when in single sweep, readjust the appropriate dela y term for the lo cation
of the problem.
V erification.
T able 3-2. Start and Stop Frequencies (GHz)
83630B /L 83640B/L 83650B /L
83623B/L
0.01 to 20
0.01 to 26.5 0.01 to 40 0.01 to 50
4.5 to 18 4 to 23 4.5 to 32 4.5 to 33
10 to 20 10 to 20 10 to 40 10 to 50
2 to 14.5 2 to 14.5 2 to 23 2 to 23
1 F or 83622B/24B only , set from 2 GHz to 20 GHz
Compare sweeps of 0 ms and 1s and also single sweeps at fast sp eed. If a p ow er drop of
0.5 division or greater is noted, readjust the appropriate calibration constant. If problems
o ccur in other bands, adjust the appropriate YTM Bx Dly T erms for that band.
Related P erformance T ests
Maxim um Lev eled P ow er
In Case Of Difficulty
oscilloscop e is scaled to 33 mV/dB.
Swept CW Gener ator T r oublesho oting Guide .

This pro cedure maximizes RF p ow er b y tracking the YIG tuned lter to the RF output
frequency . Initial tracking is done in single band sweeps at slow sweep sp eeds to eliminate
the e ects of bandcross and h ysteresis in the tuning coil. (Those will b e corrected with dela y
calibration constants.) Auto tracking is initiated to optimize the slow sweep tracking.
The YTF dela y comp ensation adjustmen ts maximize p ow er for fast single- and m ulti-band
sweeps. The YTM risetime adjustmen ts are done in m ulti-band fast sweeps to optimize p ow er
at the start of eac h frequency band.
A squegging clamp adjustmen t limits the p ow er and is adjusted for maximum output p ow er
without squegging. A \squegg" is a distortion in the RF output caused b y to o m uc h p ow er to
the ampli er/multiplier and c haracterized b y a p ow er drop out in a p ortion of the trace, or a
p ow er drop out ov er a broad frequency range.
If y ou ha v e replaced either the YO or the YO driver, rst initiate an auto
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Auto Track .) If auto tracking
passes and the instrumen t passes the \Maxim um Lev eled P ow er" p erformance
test, do not con tinue with this pro cedure. If auto tracking fails, (an error
message is display ed), con tinue with this pro cedure.

The follo wing explanations are provided for a b etter understanding of the ampli er/ lter
adjustmen ts. Refer to these explanations as often as necessary . They will help make these
adjustmen ts easier. (The adjustmen t pro cedure follo ws.)
Offset and Gain Adjustments.
F or all o set and gain adjustmen ts, adjust thr ough the bandpass. Keep adjusting un til the
p ow er p eaks and then drops o . Then reset the adjustmen t to the p eak ed p oint.
F or all o set and gain adjustmen ts, if p ow er stays p eak ed ov er sev eral calibration constant
v alues, set the calibration constant to the middle v alue of the p eak ed range.
Squegg Clamp Calibration Constants.
Squegg clamp calibration constants are av ailable to limit the p ow er in to the
ampli er/multiplier. The band 1 adjustmen t is set for a sp eci c output p ow er at the start
of band 1. All other squegg clamp adjustmen ts are set to a default v alue to allo w maxim um
output p ow er.
When increasing an y squegg clamp calibration constant, p ow er is increased to o far when an y
of the following o ccurs:
An y p ortion of the trace is distorted or p ow er drops out in a p ortion of the trace (sphere
squegging).
P ow er drops ov er a broad frequency range.
P ow er stops increasing on an y p ortion of the trace.
Adjusting for the Center of the P assband.
The passband and tuning linearity of the ampli er/ lter v aries with frequency . The
ampli er/ lter adjustmen ts track the ampli er/ lter passband with the ampli er/multiplier
output frequency ov er the full frequency range of the syn thesizer.
Adjustment Help · Agilent 8360 Adjustments · Amplifier/Filter Adjustments

Single Band Delay Compensation Adjustments.
If a drop in p ow er greater than 1.5 dB o ccurs when in single sweep, increase the A dela y
term to put the A term at the high end of the p eak.
F or example, for the following calibration constant v alues:
Start of the p eak = 1800
Middle of the p eak = 1950
End of the p eak = 2100
Set the A term to 2050
See Figure 3-16 for a graphic represen tation of the frequencies a ected b y the calibration
constants.
Setting the F astest Sw eep Time.
Setting the sweep time to 0 milliseconds sets the syn thesizer to the fastest sweep time for
the frequency range b eing swept. Eac h time the frequency range is c hanged, the sweep time
m ust b e reset to 0 ms in order to maintain the fastest sweep time.
Sw eep Speed Related Adjustments.
Delay comp ensation and risetime adjustmen ts are a ected b y sweep sp eed. The adjustmen ts
are p erformed at fast sweep sp eeds whic h are the w orst case. All other adjustmen ts are
p erformed at slow sweep sp eeds; they are not a ected b y sweep sp eed.
If y ou ha v e a sweep sp eed problem, it is probably a ected b y dela y comp ensation or risetime
adjustmen ts.
YTF Bandcross Delay T erms.
The YTF Bx Dly terms are either o set or gain terms. A1 denotes the o set term for band 1.
B1 denotes the gain term for band 1. The o set (A) should b e adjusted to maximize p ow er at
the b eginning of the band. The gain (B) is adjusted to maximize p ow er tow ard the end of the
band.
See Figure 3-16 and Figure 3-17 for a graphic represen tation of the frequencies a ected b y the
calibration constants.
3-34 Adjustments · Amplifier/Filter Adjustments


Figure 3-16. Single-Band Delay and Risetime Compensation
Agilent 8360 Adjustments · Amplifier/Filter Adjustments


Figure 3-17. Multi-Band Delay and Risetime Compensation
3-36 Adjustments · Amplifier/Filter Adjustments

v ersus B sweep capabilit y . Do not connect the p ow er meter y et. (See Figure 3-19 for
lo cation detail of XA10J2 pin 55 on the motherb oard.) P ow er on all the instrumen ts and
let them w arm up for at least one hour.
on the syn thesizer and, for the 83640B /L and 83650B /L press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Disable Doubler (asterisk on)
Zero and calibrate the p ow er meter/sensor.
Set the p ow er meter to dBm mo de.

Figure 3-18. Amplifier/Filter Adjustments Setup
Analog Oscilloscop e HP/Agilent 1740A
Oscilloscop e Prob es HP/Agilent 10431A
P ow er Meter HP/Agilent 436A/7A/8A
P ow er Sensor HP/Agilent 8485A (83620B/22B/24B,
P ow er Sensor HP/Agilent 8487A (83640B/L and 83650B /L)
A tten uator 10 dB HP/Agilent 8493C Option 010 (83620B /22B/24B,
A tten uator 10 dB HP/Agilent 8490D Option 010 (83640B /L and 83650B/L)


Figure 3-19. XA10J2 Location
calibration constants. On the syn thesizer, set:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Select Cal En ter the n um b er of the rst calibration constant from T able 3-4 and
terminate the en try with
. When using the front panel em ulator softw are,
pressing the arrow k eys will allo w y ou to v ary the calibration constants. The calibration
constant sho wn at the top of the \display" window is the activ e constant.
Modify Cal En ter the default v alue from T able 3-4 and terminate the en try with
.
Using the up/do wn arrow k eys, select eac h of the rest of the calibration constants in
T able 3-4 and set their default v alues b y en tering the v alue on the n umeric k eypad.
T erminate y our en try with the
k ey .

T able 3-4. Amplifier/Filter Calibration Constants and Default V alues
Numb er Description Default
131 Squegg Clamp 1A 255 Maximize p ow er across the
band without squegging.
132 Squegg Clamp 2A 255
Adjusted at slow sw eep sp eeds
133 Squegg Clamp 3A 255
and single sw eep mo de where the
134 Squegg Clamp 4A 255
SYTM sphere is most likely to
squegg.
135 Squegg Clamp 5A 255
136 Squegg Clamp 6A 255
137 Squegg Clamp 7A 255
205 YTM Rise; Band 1 15 Minimize p ow er drop outs at the
start of band. Adjustmen t is only
206 YTM Rise; Band 2 A1 20
e ectiv e at fast sw eep sp eeds.
207 YTM Rise; Band 2 B1 2
208 YTM Rise; Band 2 A2 15
209 YTM Rise; Band 3 A1 70
210 YTM Rise; Band 3 B1 7
211 YTM Rise; Band 3 A2 15
212 YTM Rise; Band 3 B2 15
213 YTM Rise; Band 3 A3 15
214 YTM Rise; Band 4 15
215 YTM Rise; Band 5 15
216 YTM Rise; Band 6 10
217 YTM Rise; Band 7 10

T able 3-4.
Amplifier/Filter Calibration Constants and Default V alues (continued)
Numb er Description Default
249 Autotrack D A C Setting 2048 Sets p ow er level during
Autotrack or P ow er P eaking
449 D Amp Clamp 1 255 Protect the bridge detector
from excessive p ow er when
450 D Amp Clamp 2 255
doubler amp mo de is selected.
451 D Amp Clamp 3 255
452 D Amp Clamp 4 255
453 D Amp Clamp 5 255
454 D Amp Clamp 6 255
455 D Amp Clamp 7 255
561 YTF Delay T erm A Hrm 1 1000 Maximize p ow er over rst 20%
of band. A ects fast sw eeps
562 YTF Delay T erm A Hrm 2 1000
only . Single sw eep mo de is also
563 YTF Delay T erm A Hrm 3 1000
critical.
564 YTF Delay T erm A Hrm 4 1000
565 YTF Delay T erm A Hrm 5 1000
566 YTF Delay T erm A Hrm 6 1000
567 YTF Delay T erm A Hrm 7 1000
571 YTF Delay T erm B Hrm 1 300 Maximize p ow er over higher 30%
572 YTF Delay T erm B Hrm 2 300
of band. A ects fast sw eeps only .
573 YTF Delay T erm B Hrm 3 300
574 YTF Delay T erm B Hrm 4 300
575 YTF Delay T erm B Hrm 5 300
576 YTF Delay T erm B Hrm 6 300
577 YTF Delay T erm B Hrm 7 300
591 YTF Bx Dly T erm A1 1000 Delay comp ensation for m ulti-
592 YTF Bx Dly T erm A2 1000
band sw eeps only . Adjust for
maxim um p ow er at low er 50% of
593 YTF Bx Dly T erm A3 1000
band. A ects fast sw eep sp eeds
594 YTF Bx Dly T erm A4 1000
only .
595 YTF Bx Dly T erm A5 1000
596 YTF Bx Dly T erm A6 1000
597 YTF Bx Dly T erm A7 1000

T able 3-4.
Amplifier/Filter Calibration Constants and Default V alues (continued)
Numb er Description Default
601 YTF Bx Dly T erm B1 300 Delay comp ensation for m ulti-
band sw eeps only . Adjust for
602 YTF Bx Dly T erm B2 300
maxim um p ow er at higher 30% of
603 YTF Bx Dly T erm B3 300
band. A ects fast sw eep sp eeds
only .
604 YTF Bx Dly T erm B4 300
605 YTF Bx Dly T erm B5 300
606 YTF Bx Dly T erm B6 300
607 YTF Bx Dly T erm B7 300
621 *YTF Gain Band 1 2048 Maximize p ow er over higher 10%
of band. Adjust at slow sw eep
622 *YTF Gain Band 2 2048
sp eeds.
623 *YTF Gain Band 3 2048
624 *YTF Gain Band 4 2048
625 *YTF Gain Band 5 2048
626 *YTF Gain Band 6 2048
627 *YTF Gain Band 7 2048
631 *YTF O set Band 1 2048 Maximize p ow er over low er 10%
of band. Adjust at slow sw eep
632 *YTF O set Band 2 2048
sp eeds.
633 *YTF O set Band 3 2048
634 *YTF O set Band 4 2048
635 *YTF O set Band 5 2048
636 *YTF O set Band 6 2048
637 *YTF O set Band 7 2048
Amplifier/Filter Offset and Gain Adjustment
Band 1 Adjustment.
2.0
SWEEP TIME
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Leveling Mode ALCoff (asterisk on)
POWER LEVEL
Channel A:

Adjust Channel A o set, v olts/division, and horizontal p osition con trols for a trace that
lls the full horizontal display . Adjust the v ertical p osition with the A v ertical p osition
con trol.
In the following adjustmen t p ow er drop outs may o ccur due to squegging.
Ignore the squegging (the squegg clamp will b e adjusted later) and p eak the
adjustmen t for maximum p ow er.
b egins to drop.
divisions).
Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace
shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 7 through 10 un til it
is. Then c hange calibration constant #631 bac k to the recorded v alue.
across the band. The clamp is adjusted correctly when the follo wing is true:
a. An increase in the calibration constant v alue decreases the output p ow er.
b. A small decrease in the calibration constant v alue results in no c hange in output
p ow er.
an 83623L, con tinue with the next step. Otherwise, adjust calibration constant #131,
Squegg Clamp 1A as describ ed here, to protect the bridge detector from excessiv e p ow er:
a. Set the oscilloscop e v ertical p osition as necessary to display the trace.
b. On the syn thesizer, set:
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Manual Sweep (asterisk on)
c. Using the rotary knob, nd the p eak of the sweep as display ed b y the oscilloscop e.
d. If the p ow er lev el of the p eak p oint is greater than 21.5 dBm for 83620B /22B and
83630B /L or 23.5 dBm for 83640B /L and 83650B /L, as display ed on the p ow er meter,
decrease Squegg Clamp 1A un til the p ow er lev el is less than these v alues.
If the p ow er lev el of the p eak p oint is already less than these v alues, return the
syn thesizer to con tinuous sweep and decrease Squegg Clamp 1A just un til the p eak
p oint b egins to drop (this indicates that p ow er to the bridge detector is clamp ed).
here. Otherwise, con tinue with the next step.
a. Set the syn thesizer's sweep time to 500 ms.
b. Adjust the calibration constant #131, Squegg Clamp 1A for maximum p ow er without
squegging.
c. Chec k for squegging b y setting the syn thesizer's sweep time to 0 ms. If the p ow er
increase is greater 2 dB in an y part of the band, then squegging is o ccurring.

d. If squegging o ccurs, readjust calibration constant #131 so that the p ow er di erence is
2 dB or less from a 500 ms (slo w) sweep to a 0 ms (fast) sweep.
Adjustmen t." Otherwise, adjust calibration constant #449, DAmp Clamp 1 as describ ed
here, to protect the bridge detector from excessiv e p ow er while in doubler amp mo de:
a. On the syn thesizer, set:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Disable Doubler (asterisk o )
PO WER
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Doubler Amp Mode On (asterisk on)
b. Set the oscilloscop e v ertical p osition as necessary to display the trace.
c. On the syn thesizer, set:
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Manual Sweep (asterisk on)
d. Using the syn thesizer rotary knob, nd the p eak of the sweep as display ed b y the
oscilloscop e.
e. If the p ow er lev el of the p eak p oint is greater than 23.5 dBm, as display ed on the
p ow er meter, decrease DAmp Clamp 1 un til the p ow er lev el is less than this lev el.
If the p ow er lev el of the p eak p oint is already less than this lev el, set the syn thesizer
bac k to con tinuous sweep and decrease DAmp Clamp 1 just un til the p eak p oint
b egins to drop (this indicates that p ow er to the bridge detector is clamp ed).
sweep if y ou are still in man ual sweep.
Band 2 Adjustment.
STOP
13.5
b egins to drop.
Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace
shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 19 through 22 un til
it is. Then c hange calibration constant #632 bac k to the recorded v alue.
across the band.
an 83623L, con tinue with the next step. Otherwise, adjust calibration constant #132,
Squegg Clamp 2A as describ ed here, to protect the bridge detector from excessiv e p ow er:
a. Set the oscilloscop e v ertical p osition as necessary to display the trace.
b. On the syn thesizer, set:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Manual Sweep (asterisk on)
c. Using the rotary knob, nd the p eak of the sweep as display ed b y the oscilloscop e.

d. If the p ow er lev el of the p eak p oint is greater than 20 dBm for 83620B /22B and
83630B /L or 22.7 dBm for 83640B /L and 83650B /L, as display ed on the p ow er meter,
decrease Squegg Clamp 2A un til the p ow er lev el is less than these v alues.
If the p ow er lev el of the p eak p oint is already less than these v alues, return the
syn thesizer to con tinuous sweep and decrease Squegg Clamp 2A just un til the p eak
p oint b egins to drop (this indicates that p ow er to the bridge detector is clamp ed).
here. Otherwise, con tinue with the next step.
a. Set the syn thesizer's sweep time to 500 ms.
b. Adjust the calibration constant #132, Squegg Clamp 2A for maximum p ow er without
squegging.
c. Chec k for squegging b y setting the syn thesizer's sweep time to 0 ms. If the p ow er
increase is greater 2 dB in an y part of the band, then squegging is o ccurring.
d. If squegging o ccurs, readjust calibration constant #132 so that the p ow er di erence is
2 dB or less from a 500 ms (slow) sweep to a 0 ms (fast) sweep.
Adjustmen t." Otherwise, adjust calibration constant #450, DAmp Clamp 2 as describ ed
here, to protect the bridge detector from excessiv e p ow er while in doubler amp mo de:
a. On the syn thesizer, set:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Disable Doubler (asterisk o )
PO WER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Doubler Amp Mode On (asterisk on)
b. Set the oscilloscop e v ertical p osition as necessary to display the trace.
c. On the syn thesizer, set:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Manual Sweep (asterisk on)
d. Using the syn thesizer rotary knob, nd the p eak of the sweep as display ed b y the
oscilloscop e.
e. If the p ow er lev el of the p eak p oint is greater than 22.7 dBm, as display ed on the
p ow er meter, decrease DAmp Clamp 2 un til the p ow er lev el is less than these v alues.
If the p ow er lev el of the p eak p oint is already less than these v alues, set the
syn thesizer bac k to con tinuous sweep and decrease DAmp Clamp 2 just un til the p eak
p oint b egins to drop (this indicates that p ow er to the bridge detector is clamp ed).
sweep if y ou are still in man ual sweep.
Band 3 Adjustment.
13.5
POWER LEVEL
b egins to drop.
There are tw o sets of p oten tiometers on the A12 assembly with iden tical
names. Make c ertain y ou are adjusting the correct ones.


Figure 3-20. A12 Assembly P otentiometer Locations
(3 to 10 divisions).
Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace
shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 31 through 35 un til
it is. Then c hange calibration constant #633 bac k to the recorded v alue.
across the band.
adjust calibration constant #133, Squegg Clamp 3A as describ ed here, to protect the
bridge detector from excessiv e p ow er:
a. Set the oscilloscop e v ertical p osition as necessary to display the trace.
b. On the syn thesizer, set:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Manual Sweep (asterisk on)
c. Using the rotary knob, nd the p eak of the sweep as display ed b y the oscilloscop e.
d. If the p ow er lev el of the p eak p oint is greater than 21.5 dBm for 83620B /22B and
83630B /L or 22.3 dBm for 83640B /L and 83650B /L, as display ed on the p ow er meter,
decrease Squegg Clamp 3A un til the p ow er lev el is less than these v alues.
If the p ow er lev el of the p eak p oint is already less than these v alues, return the
syn thesizer to con tinuous sweep and decrease Squegg Clamp 3A just un til the p eak
p oint b egins to drop
(this indicates that p ow er to the bridge detector is clamp ed).

Adjustmen t." Otherwise, adjust calibration constant #451, DAmp Clamp 3 as describ ed
here, to protect the bridge detector from excessiv e p ow er while in doubler amp mo de:
a. On the syn thesizer, set:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SERVICE
Disable Doubler (asterisk o )
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PO WER
MENU
Doubler Amp Mode On (asterisk on)
b. Set the oscilloscop e v ertical p osition as necessary to display the trace.
c. On the syn thesizer, set:
SWEEP
MENU
Manual Sweep (asterisk on)
d. Using the syn thesizer rotary knob, nd the p eak of the sweep as display ed b y the
oscilloscop e.
e. If the p ow er lev el of the p eak p oint is greater than 22.3 dBm, as display ed on the
p ow er meter, decrease DAmp Clamp 3 un til the p ow er lev el is less than these v alues.
If the p ow er lev el of the p eak p oint is already less than these v alues, set the
syn thesizer bac k to con tinuous sweep and decrease DAmp Clamp 3 just un til the p eak
p oint b egins to drop (this indicates that p ow er to the bridge detector is clamp ed).
sweep if y ou are still in man ual sweep.
Band 4 Adjustment (83630B/L, 83640B/L, and 83650B/L only).
26.5
( 83630B /L and 83650B /L )
25.5
STOP
POWER LEVEL
b egins to drop.
A12R110 S2
A12R105 B4BP1
A12R106 BP2
division on the display .
If there is no noticeable increase in p ow er in the next tw o adjustmen ts, leav e
the p oten tiometers at their preset p osition (fully clo c kwise).
divisions. Stop when the p ow er b egins to decrease.

Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace
shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 46 through 51 un til
it is. Then c hange calibration constant #634 bac k to the recorded v alue.
across the band.
here, to protect the bridge detector from excessiv e p ow er:
a. Set the oscilloscop e v ertical p osition as necessary to display the trace.
b. On the syn thesizer, set:
SWEEP
MENU
Manual Sweep (asterisk on)
c. Using the rotary knob, nd the p eak of the sweep as display ed b y the oscilloscop e.
d. If the p ow er lev el of the p eak p oint is greater than 18.9 dBm, as display ed on the
p ow er meter, decrease Squegg Clamp 4A un til the p ow er lev el is less than these
v alues.
If the p ow er lev el of the p eak p oint is already less than these v alues, return the
syn thesizer to con tinuous sweep and decrease Squegg Clamp 4A just un til the p eak
p oint b egins to drop
(this indicates that p ow er to the bridge detector is clamp ed).
DAmp Clamp 4 as describ ed here, to protect the bridge detector from excessiv e p ow er
while in doubler amp mo de:
a. On the syn thesizer, set:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Disable Doubler (asterisk o )
PO WER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Doubler Amp Mode On (asterisk on)
b. Set the oscilloscop e v ertical p osition as necessary to display the trace.
c. On the syn thesizer, set:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Manual Sweep (asterisk on)
d. Using the syn thesizer rotary knob, nd the p eak of the sweep as display ed b y the
oscilloscop e.
e. If the p ow er lev el of the p eak p oint is greater than 23.5 dBm, as display ed on the
p ow er meter, decrease DAmp Clamp 4 un til the p ow er lev el is less than these v alues.
If the p ow er lev el of the p eak p oint is already less than these v alues, set the
syn thesizer bac k to con tinuous sweep and decrease DAmp Clamp 4 just un til the p eak
p oint b egins to drop (this indicates that p ow er to the bridge detector is clamp ed).
sweep if y ou are still in man ual sweep.
Band 5 Adjustment (83640B/L and 83650B/L only).
25.5
ST ART
26.5
STOP
( 83640B /L)
33.5
STOP
POWER LEVEL
b egins to drop.

Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace
shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 61 through 63 un til
it is. Then c hange calibration constant #635 bac k to the recorded v alue.
across the band.
Band 6 Adjustment (83640B/L and 83650B/L only).
ST ART
ST ART
33.5
STOP
( 83640B /L)
STOP
( 83650B /L)
POWER LEVEL
b egins to drop.
Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace
shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 70 through 72 un til
it is. Then c hange calibration constant #636 bac k to the recorded v alue.
across the band.
Band 7 Adjustment (83650B/L only).
STOP
POWER LEVEL
b egins to drop.
Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace
shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 79 through 81 un til
it is. Then c hange calibration constant #637 bac k to the recorded v alue.

across the band.
Auto Tracking V erification.
#621 YTF Gain; Band 1
#622 YTF Gain; Band 2
#623 YTF Gain; Band 3
#624 YTF Gain; Band 4
#625 YTF Gain; Band 5
#626 YTF Gain; Band 6
#627 YTF Gain; Band 7
#631 YTF O set; Band 1
#632 YTF O set; Band 2
#633 YTF O set; Band 3
#634 YTF O set; Band 4
#635 YTF O set; Band 5
#636 YTF O set; Band 6
#637 YTF O set; Band 7
#225 YTM Gain; Band 1
#226 YTM Gain; Band 2
#227 YTM Gain; Band 3
#228 YTM Gain; Band 4
#229 YTM Gain; Band 5
#230 YTM Gain; Band 6
#231 YTM Gain; Band 7
#238 YTM O set; Band 1
#239 YTM O set; Band 2
#240 YTM O set; Band 3
#241 YTM O set; Band 4
#242 YTM O set; Band 5
#243 YTM O set; Band 6
#244 YTM O set; Band 7
go o d 50 imp edance match suc h as a 10 dB attenuator or a p ow er sensor (not required
for syn thesizers with a step attenuator). Press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
USER CAL
W ait for the syn thesizer to complete auto tracking. If auto tracking fails (an error
message is display ed) rep eat the gain and o set adjustmen ts for the failed frequency .
2.0
CONT
SWEEP TIME
1.5
ALC
Leveling Point Intrnl (asterisk on)

F or the 83640B /L and 83650B /L only , also press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Disable Doubler (asterisk o )
Connect the p ow er meter to the syn thesizer RF OUTPUT through the p ow er sensor and
10 dB attenuator (see Figure 3-18). Adjust the oscilloscop e horizontal con trol to place
the low p ow er p oint on a v ertical graticule and note the p osition.
Select man ual sweep. Press SWEEP
Manual Sweep (asterisk on). Use the arrow
k eys to adjust the CW frequency to the minimum p ow er p oint on the oscilloscop e.
On the p ow er meter, c hec k that the p ow er at this p oint is not less than the sp eci ed
maximum lev eled p ow er (note that p ow er is attenuated b y 10 dB ). If it is, note the lev el
and frequency , and c hec k the follo wing:
a. Compare the man ually adjusted YTF and YTM gain and o set calibration constants
(step 85) to the auto track ed calibration constants. A large di erence in v alues
(several h undred coun ts) indicates a p ossible misadjustment. The auto tracking v alues
are correct.
b. If the band 3 v alues are within 500 coun ts of the limits of their ranges, rep eat the
B3S1, B4S1, and B4BP1 adjustmen ts with the doubler disabled.
c. If the band 4 v alues are within 500 coun ts of the limits of the range, rep eat the B4S1
and B4BP1 adjustmen ts with the doubler disabled (unless y ou ha v e just readjusted
these adjustmen ts in step b).
d. If the problem con tinues, it is most likely hardw are-related. Refer to the A gilent
T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series Swept CW
Gener ator T r oublesho oting Guide .
listed in T able 3-5 and a sweep time of 1.5 seconds.
T able 3-5. Frequency Ranges
20 to 26.5 GHz 20 to 25.5 GHz 20 to 26.5 GHz
25.5 to 40 GHz 26.5 to 38 GHz
38 to 50 GHz
2.0
STOP
CONT
SWEEP TIME
1.5
oscillations or lev el squegging.
a. The SRD bias A or B calibration constants for the a ected band might eliminate
oscillation. The problem is with the A term if the oscillations are near the b eginning
of the band. It is with the B term if the oscillations are near the end of the band.
SRD bias has a range. It can b e ov erbiased whic h causes lev el squegging. If so,
decrease the v alue of the SRD bias calibration constants and rep eat the adjustmen t.
b. If the SRD bias calibration constants ha v e no e ect, c hec k the ALC Mo d Gain
calibration constant for the appropriate band.
listed in T able 3-5 and a sweep time of 1.5 seconds.

5 GHz to 50 GHz or 10 GHz to 45 GHz.
Set the sweep times to greater than 200 ms for single-band sweeps and less than 1.5
seconds for m ulti-band sweeps since dela y comp ensation has not y et b een adjusted.
F or eac h frequency range set, compare the signal lev els of single sweep to con tinuous
sweep. Band frequency ranges are as follo ws:
83650B /L
0 0.01 to < 2.0 0.01 to < 2.0 0.01 to < 2.0
1 2.0 to 7.0 2.0 to 7.0 2.0 to 7.0
2 7.0 to < 13.5 7.0 to < 13.5 7.0 to < 13.5
3 13.5 to 20.0 13.5 to 20.0 13.5 to 20.0
4 > 20.0 to 26.5 > 20.0 to 25.5 > 20.0 to 26.5
5 25.5 to < 32 26.5 to < 33.5
6 32 to < 40 33.5 to < 38
7 38 to 50
Problems here are typically asso ciated with sphere heating while aw aiting the start of sweep
trigger. If necessary adjust the squegging clamp calibration constant for the appropriate band.
Single-Band Amplifier/Filter Delay .
2.0
STOP
CONT
SWEEP TIME
POWER LEVEL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Leveling Mode ALCoff (asterisk on)
F or the 83640B /L and 83650B /L only , also set:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Disable Doubler (asterisk on)
b egins to drop.

The oscilloscop e v ertical sensitivit y is initially set to 50 mV/division (5
mV/division with a 10:1 prob e) providing a logarithmic scale of 1.5
dB/division. If necessary , decrease the v ertical sensitivity for some frequency
bands to 100 mV/division in order to display the en tire trace on the CR T
(the oscilloscop e is already set to 10 mV/division for the 83623B /L and the
83624B ). If y ou do c hange the v ertical sensitivit y b y half, remem b er that the
logarithmic scale m ust c hange b y half also.
Band 1 Adjustment.
p ow er c hanges less than 2/3 division ( 1 dB.) If not, adjust the appropriate rise or dela y
term (steps 98 through 100) and rep eat this step.
If fast sweep has more p ow er than slow sweep do not readjust. Con tinue with
the next step.
sev eral times to initiate sev eral sweeps and c hec k that a drop in p ow er 1
division o ccurs when in single sweep. If it is less than 1 division, skip the remainder of
\Band 1 Adjustmen t" and con tinue with \Band 2 Adjustmen t."
If the drop in p ow er is > 1 division, increase the A term of the dela y adjustmen t #561 to
put the A term at the high end of the p eak.
Then rep eat from step 95.
low end of band 1. Set the calibration constant for the middle of the p eak ed range.
high end of band 1. Set the calibration constant for the middle of the p eak ed range.
start of band 1.
Band 2 Adjustment.
If the oscilloscop e is not already set to 5 mV/division
(or 10 mV/division for the 83623B /L and the 83624B ), do so no w. The
logarithmic scale is calculated with the 5 mV/division v alue so for mo dels
using 10 mV/division the logarithmic scale m ust c hange b y half.
13.5
CONT
SWEEP TIME
POWER LEVEL
b egins to drop.
p ow er c hanges less than 2/3 division. If not, adjust the appropriate rise or delay term
(steps 106 through 108) and then rep eat this step.

sev eral times to initiate sev eral sweeps. Note an y drop in p ow er.
b. Press
RF ON/OFF
. Then press
again (LED on) and note an y drop
in p ow er.
c. The p ow er loss noted in step a can b e no smaller than the p ow er loss noted in step b.
If the step a p ow er loss is greater than the step b p ow er loss b y 1 division ( 1.5 dB),
increase the A term of the dela y adjustmen t #562 to put the A term at the high end
of the p eak. Then rep eat from step 103.
If the p ow er loss in step a is not smaller than that in step b, skip the rest of \Band 2
Adjustmen t" and con tinue with \Band 3 Adjustmen t."
low end of band 2.
high end of band 2.
the start of band 2.
Band 3 Adjustment.
13.5
STOP
CONT
SWEEP TIME
POWER LEVEL
b egins to drop.
p ow er c hanges less than 2/3 division ( 1 dB). If not, adjust the appropriate rise or dela y
terms (steps 114 through 116) and then rep eat this step.
sev eral times to initiate sev eral sweeps. Note an y drop in p ow er.
b. Press
. Then press
again (LED on) and note an y drop
RF ON/OFF
RF ON/OFF
in p ow er.
c. The p ow er loss noted in step a can b e no smaller than the p ow er loss noted in step b.
If the step a p ow er loss is greater than the step b p ow er loss b y 1 division ( 1.5 dB),
increase the A term of the dela y adjustmen t #573 to put the A term at the high end
of the p eak. Then rep eat from step 111.
If the p ow er loss in step a is not smaller than that in step b, skip the rest of \Band 3
Adjustmen t" and con tinue with \Band 4 Adjustmen t."
low end of band 3.
high end of band 3.
the start of band 3.

Band 4 Adjustment (83630B/L, 83640B/L, and 83650B/L).
26.5
( 83630B /L and 83650B /L )
STOP
25.5
CONT
SWEEP TIME
POWER LEVEL
b egins to drop.
p ow er c hanges less than 2/3 division. If not, adjust the appropriate rise term (steps
122 and 123) and then rep eat this step.
sev eral times to initiate sev eral sweeps. Note an y drop in p ow er.
b. Press
. Then press
again (LED on) and note an y drop
RF ON/OFF
RF ON/OFF
in p ow er.
c. The p ow er loss noted in step a can b e no smaller than the p ow er loss noted in step b.
If the step a p ow er loss is greater than the step b p ow er loss b y 1 division ( 1.5 dB),
increase the A term of the dela y adjustmen t #564 to put the A term at the high end
of the p eak. Then rep eat from step 119.
If the p ow er loss in step a is not smaller than that in step b, skip the rest of \Band 4
Adjustmen t" and con tinue with \Band 5 Adjustmen t."
low end of band 4.
high end of band 4.
Band 5 Adjustment (83640B/L and 83650B/L only).
25.5
ST ART
26.5
STOP
( 83640B /L)
33.5
STOP
CONT
SWEEP TIME
POWER LEVEL
b egins to drop.
p ow er c hanges less than 2/3 division ( 1 dB). If not, adjust the appropriate rise or delay
term (steps 129 through 131) and then rep eat this step.
sev eral times to initiate sev eral sweeps. Note an y drop in p ow er.
b. Press
RF ON/OFF
. Then press
again (LED on) and note an y drop
in p ow er.
c. The p ow er loss noted in step a can b e no smaller than the p ow er loss noted in step b.
If the step a p ow er loss is greater than the step b p ow er loss b y 1 division ( 1.5 dB),
increase the A term of the delay adjustmen t #565 to put the A term at the high end
of the p eak. Then rep eat from step 126.

If the p ow er loss in step a is not smaller than that in step b, skip the rest of \Band 5
Adjustmen t" and con tinue with \Band 6 Adjustmen t."
low end of band 5.
high end of band 5.
start of band 5.
Band 6 Adjustment (83640B/L and 83650B/L only).
( 83640B /L)
ST ART
33.5
STOP
( 83640B /L)
STOP
( 83650B /L)
CONT
SWEEP TIME
POWER LEVEL
b egins to drop.
p ow er c hanges less than 2/3 division ( 1 dB). If not, adjust the appropriate rise or dela y
term (steps 137 through 139) and then rep eat this step.
sev eral times to initiate sev eral sweeps. Note an y drop in p ow er.
b. Press
RF ON/OFF
. Then press
again (LED on) and note an y drop
in p ow er.
c. The p ow er loss noted in step a can b e no smaller than the p ow er loss noted in step b.
If the step a p ow er loss is greater than the step b p ow er loss b y 1 division ( 1.5 dB),
increase the A term of the dela y adjustmen t #566 to put the A term at the high end
of the p eak. Then rep eat from step 134.
If the p ow er loss in step a is not smaller than that in step b, skip the rest of \Band 6
Adjustmen t" and con tinue with \Band 7 Adjustmen t."
low end of band 6.
high end of band 6.
start of band 6.
Band 7 Adjustment (83650B/L only).
STOP
CONT
SWEEP TIME
POWER LEVEL
b egins to drop.

p ow er c hanges less than 2/3 division ( 1 dB). If not, adjust the appropriate rise or dela y
term (steps 145 through 147) and then rep eat this step.
sev eral times to initiate sev eral sweeps. Note an y drop in p ow er.
b. Press
. Then press
again (LED on) and note an y drop
RF ON/OFF
RF ON/OFF
in p ow er.
c. The p ow er loss noted in step a can b e no smaller than the p ow er loss noted in step b.
If the step a p ow er loss is greater than the step b p ow er loss b y 1 division ( 1.5 dB),
increase the A term of the dela y adjustmen t #567 to put the A term at the high end
of the p eak. Then rep eat from step 142.
If the p ow er loss in step a is not smaller than that in step b, skip the rest of \Band 7
Adjustmen t" and con tinue with \Multi-band YTF Delay ."
low end of band 7.
high end of band 7.
start of band 7.
Multi-band YTF Delay .
STOP
CONT
SWEEP TIME
POWER LEVEL
b egins to drop.
Syn thesizer in tensity markers are set to iden tify the bandcross frequencies (7.0,
13.5 GHz, and 20.0 GHz). Reduce oscilloscop e trace in tensity to iden tify these
markers. Y ou can also iden tify bandcrosses b y temp orarily removing the cable
from the Z-AXIS BLANK/MKRS connector on the syn thesizer rear panel.
Make sure this cable is connected when making adjustmen ts.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
13.5
end of band 3.
end of band 3.
the end of band 3.
p ow er c hanges less than 2/3 division ( 1 dB). If the c hange in p ow er is greater than 2/3
division, readjust the A term of the delay adjustmen t #593 for low end problems, or the
B term of the dela y adjustmen t #603 for high end problems.

SWEEP TIME
sev eral times to initiate sev eral sweeps. If a drop in p ow er greater than 1
division o ccurs when in single sweep, increase the A term of the dela y adjustmen t #593
to put the A term at the high end of the p eak.
CONT
SWEEP TIME
POWER LEVEL
b egins to drop.
p ow er c hanges less than 2/3 division ( 1 dB) across b oth bands. If the c hange is greater
than 2/3 division, readjust the appropriate dela y term.
of band 2 (start of sweep).
of band 2.
the start of band 3.
POWER LEVEL
b egins to drop.
the start of band 3.
sev eral times to initiate sev eral sweeps. If a drop in p ow er greater than 1
division o ccurs when in single sweep, readjust the appropriate dela y term for the lo cation
of the problem.
2.0
CONT
SWEEP TIME
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
POWER LEVEL
b egins to drop.
of band 1.
end of band 1.
the start of band 2.
the start of band 3.
c hanges less than 2/3 division ( 1 dB) across all bands. If the c hange is greater than 2/3
division, readjust the appropriate delay term.

sev eral times to initiate sev eral sweeps. If a drop in p ow er greater than 1
division o ccurs when in single sweep, readjust the appropriate dela y or risetime term for
the lo cation of the problem.
POWER LEVEL
b egins to drop.
the end of band 3.
the end of band 2.
Band 4 Multi-band YTF Delay (83630B/L, 83640B/L, and 83650B/L only).
2.0
26.5
( 83630B /L and 83650B /L )
25.5
STOP
CONT
SWEEP TIME
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER
POWER LEVEL
b egins to drop.
of band 4.
of band 4.
start of band 4.
p ow er c hanges less than 2/3 division ( 1 dB) across b oth bands. If the c hange is greater
than 2/3 division, readjust the appropriate dela y term.
sev eral times to initiate sev eral sweeps. If a drop in p ow er greater than 1
division o ccurs when in single sweep, readjust the appropriate delay term for the lo cation
of the problem.
CONT
SWEEP TIME
POWER LEVEL
b egins to drop.
sev eral times to initiate sev eral sweeps. If a drop in p ow er greater than 1/3
division o ccurs when in single sweep, readjust the appropriate delay or risetime term for
the lo cation of the problem.

Operation Check.
Start and Stop Frequencies (GHz)
83622B /24B 83630B /L 83640B/L 83650B /L
83623B/L
0.045 to 20 2 to 20 0.045 to 26.5 0.045 to 40 0.045 to 50
4.5 to 18 4.5 to 18 4.5 to 18 4.5 to 32 4.5 to 33
10 to 20 10 to 20 10 to 20 10 to 23 10 to 45
2 to 14.5 2 to 14.5 2 to 14.5 2 to 23 2 to 23
Compare sweeps of 0 ms and 1 s and also single sweeps at fast sp eed. If a p ow er drop of
0.5 division or greater is noted, readjust the appropriate calibration constant.
to store the calibration constants as protected data (in EEPR OM).
Related P erformance T ests
Maximum Lev eled P ow er
In Case Of Difficulty
oscilloscop e is scaled to 33 mV/dB.
Swept CW Gener ator T r oublesho oting Guide .
Agilent 8360 Adjustments · Amplifier/Filter Adjustments

The syn thesizer is set to 10 GHz. Then the A9 pulse b oard MIN adjustmen t sets the source
v oltage of the FET for the prop er SRD bias lev el.
up for at least one hour.

Figure 3-21. Lo w P o w er SRD Bias Adjustment Setup
.
Lo w P o w er SRD Bias


Figure 3-22. A9R105 Location
Related P erformance T ests
In Case of Difficulty
Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series Swept
CW Gener ator T r oublesho oting Guide .
Lo w P o w er SRD Bias
None · Agilent 8360 Adjustments

In this pro cedure, the ALC mo dulation o set and gain calibration constants are adjusted
to linearize the ALC mo dulator resp onse to the ALC p ow er lev el reference v oltage. Default
v alues are en tered for the mo dulator o set calibration constants and in ternal rm w are is
activ ated to set the mo dulator gain calibration constants. The syn thesizer is then set for a
p ow er sweep across the en tire lev eled ALC range ( 20 to the maximum lev eled p ow er). The
in tegrator lev el signal on the ALC b oard is monitored to v erify linearity .
If necessary , the mo dulator o set v alues are mo di ed.
instrumen ts and let them w arm up for at least one hour. F or 8360 L-Series, start this test
at step 2.

Figure 3-23. Modulator Offset and Gain Adjustment Setup
Digitizing Oscilloscop e HP/Agilent 54111D
Oscilloscop e Prob es HP/Agilent 10431A
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SERVICE
(A38), set the ALC o set calibration constants to the default v alues given in T able 3-6 .
Otherwise, skip to step 7 for 8360 B-Series or skip to the next adjustmen t for 8360
L-Series.
Select Cal . Using the n umeric k eypad, en ter the n um b er of the rst calibration
constant in T able 3-6 . T erminate y our en try with the
k ey .
Modulator Offset and Gain

Modify Cal . Use the n umeric k eypad to en ter the corresp onding default v alue in
Figure 3-23 . T erminate y our en try with the
k ey .
T able 3-6 and set their default v alues b y en tering the v alue on the n umeric k eypad.
T erminate y our en try with the
k ey .
T able 3-6. ALC Offset Calibration Constant Default V alues
Description Default V alue
315 ALC Mo d Ofs; Band 0 100
316 ALC Mo d Ofs; Band 1 100
317 ALC Mo d Ofs; Band 2 100
318 ALC Mo d Ofs; Band 3 100
319 ALC Mo d Ofs; Band 4 100
320 ALC Mo d Ofs; Band 5 150
321 ALC Mo d Ofs; Band 6 150
322 ALC Mo d Ofs; Band 7 150
Modulator Offset and Gain

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SERVICE
The syn thesizer will measure, calculate, and store the mo dulator gain calibration
constants.
When the adjustmen t is completed the following message app ears on the display:
Modulator Gain Cal Completed
Calibration constants were modified.
to store the calibration constants as protected data (in EEPR OM).
8360 L-Series: This adjustmen t is complete; con tinue with next adjustmen t.
8360 B-Series: Con tinue with this adjustmen t.
Channel 1:
Input Impedance 1 M
Channel 2:
Input Impedance 1 M
Timebase:
Delay Reference At left
Trigger:
Trigger Source Channel 2
Trigger Level 2 V
Display:
Number of Averages 2
Modulator Offset and Gain

SP AN
PO WER
MENU
No w set the syn thesizer for a p ow er sweep of its full ALC range. F or example, if y our
syn thesizer's sp eci ed maxim um lev eled p ow er at this fr e quency is +10 dBm then set the
p ow er sweep to 30 dB/sw eep ( 20 dBm to +10 dBm = 30 dB). Press:
MENU
POWER SWEEP
USER CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Peak RF Always (asterisk on)
display . This signal is low during the p ow er sweep.
during the p ow er sweep (while Z-axis is low) as follo ws:
Adjust the calibration constant so that the trace in the last three (farthest right)
graticules is as at as p ossible. Initiate an AM bandwidth calibration, press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AM BW Cal Once .
Repeat the adjustmen t and calibration un til the last three graticules are as at as
p ossible.
as appropriate to sweep the full ALC range. P erform an AM bandwidth calibration and
then rep eat step 14 for eac h corresp onding calibration constant.
T able 3-7. Center Frequencies and Calibration Constants
Cen ter F requencies
5.0 #316 Band 1
9.0 #317 Band 2
19.0 #318 Band 3
1 Disregard if b ey ond your synthesizer's capability .
Modulator Offset and Gain

Related P erformance T ests
Maximum Lev eled P ow er AM Bandwidth
In Case of Difficulty
Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series Swept
CW Gener ator T r oublesho oting Guide .
Modulator Offset and Gain
3-66 Adjustments

This adjustmen t sets the absolute p ow er accuracy in eac h frequency band at a CW frequency .
The digital ALC calibration arra y is loaded with zeros for the 0 dB attenuator setting to
eliminate an y p ow er o set from this source. Calibration constants set the p ow er accuracy at
four p ow er lev els.
A p oten tiometer sets the p ow er accuracy at an 18 dBm lev el in the low band
(not applicable for all mo dels).
This adjustmen t pro cedure zeros the digital ALC calibration arra y at the 0 dB
attenuator setting. The \Pow er Flatness" adjustmen t m ust b e p erformed after
this pro cedure.
Do not connect the p ow er sensor to the syn thesizer RF OUTPUT y et.

Figure 3-24. ALC P o w er Lev el Accuracy Adjustment Setup
P ow er Meter HP/Agilent 436A/7A/8A
P ow er Sensor HP/Agilent 8485A
(83620B/22B/24B, 83623B/L, and 83630B/L)
P ow er Sensor HP/Agilent 8487A (83640B/L and 83650B/L)
Connect the p ow er sensor to the syn thesizer's RF OUTPUT.
Set the CAL F A CTOR % for the p ow er sensor 1 GHz calibration factor.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RF Peak Always (asterisk on)
PO WER
MENU
Uncoupl Atten (Option 001 only)
ALC P o w er Lev el Accuracy

frequency ranges:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SERVICE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AtnStep To Cal
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Cal Freq Range Lo Band
ALC Cal Array (Atn, Freq) Zero
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Cal Freq Range Hi Band
ALC Cal Array (Atn, Freq) Zero
The default v alues given in the next step are a general starting p oint. If the
curren t ALC p ow er accuracy is within 1 dB, few er rep etitions are required if
y ou start with the existing calibration constants unc hanged.
constants. On the syn thesizer, set:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SERVICE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Select Cal En ter the n um b er of the calibration constant from T able 3-8 and terminate
the en try with
.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Modify Cal En ter the default v alue from T able 3-8 and terminate the en try with
.
T able 3-8. ALC Calibration Constants and Default V alues
ALC Calibration Constan t Default V alue
Low Band 0.01 to 2.0 GHz
#265 L VL DA C Ofs Lo Bnd 120
#252 L VL DA C Gain Lo Bnd 20
#284 ALC Det Ofs; Lo 142
#293 ALC Log Brkpt; Lo 140
High Band 2.0 to 20 GHz
#264 L VL DA C Ofs Hi Bnd 0
#251 L VL DA C Gain Hi Bnd 20
#283 ALC Det Ofs; Hi 134
#292 ALC Log Brkpt; Hi 131
1 Disregard if b eyond your synthesizer's capability .
Lo w Band ALC P o w er Accuracy Adjustment
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SERVICE
First p ow er lev el in T able 3-9
ENTER
ALC P o w er Lev el Accuracy

On the syn thesizer, select
Modify Cal and, using the rotary knob, mo dify the calibration
constant so that the p ow er meter and the p ow er lev el setting are the same.
T able 3-9.
T able 3-9. P o w er Lev el and Calibration Constant Adjustment
P ow er Lev el
Calibration Constan t Adjustmen t
0.0 #265 L VL DA C Ofs Lo Bnd
10.0 #252 L VL DA C Gain Lo Bnd
20.0 #284 ALC Det Ofs; Lo
+10.0 #293 ALC Det Log Brkpt; Lo
Adjust A10R125, +20 dBm, un til the p ow er meter and the p ow er lev el setting are the
same. Figure 3-25 sho ws the lo cation of A10R125.

Figure 3-25. A10R125 Location
0.01 dBm.
High Band P o w er Accuracy Adjustment
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SERVICE
ALC P o w er Lev el Accuracy

factor.
First p ow er lev el in T able 3-10
ENTER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
On the syn thesizer, select
Modify Cal and, using the rotary knob, mo dify the calibration
constant so that the p ow er meter and the p ow er lev el setting are the same.
T able 3-10.
within 0.01 dBm.
T able 3-10. P o w er Lev el and Calibration Constant Adjustment
P ow er Lev el
Calibration Constan t Adjustmen t
0.0 #264 L VL DA C Ofs Hi Bnd
10.0 #251 L VL DA C Gain Hi Bnd
20.0 #283 ALC Det Ofs; Hi
+10.0 #292 ALC Det Log Brkpt; Hi
this man ual to store the calibration constants as protected data (in EEPR OM).
Related P erformance T ests
In Case of Difficulty
Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series Swept
CW Gener ator T r oublesho oting Guide .
ALC P o w er Lev el Accuracy

(The automated p ow er atness adjustmen t in Chapter 5 may b e used in place of this
pro cedure.)
In this pro cedure, the syn thesizer measures and corrects p ow er atness. The syn thesizer
con trols the p ow er meter via GPIB while the p ow er meter is measuring the RF output.
F or eac h syn thesizer frequency band and a 0 dB attenuator setting, con trol is given to the
syn thesizer to measure and correct p ow er atness.
This adjustmen t requires an HP/Agilent 437B P ow er Meter. The correct
p ow er sensor calibration factors m ust b e loaded and selected. This pro cedure
cannot b e run with a con troller on the GPIB, nor can it b e run from a front
panel em ulator.
syn thesizer RF output y et. Let b oth instrumen ts w arm up for at least one hour.

Figure 3-26. P o w er Flatness Adjustment Setup
P ow er Meter HP/Agilent 437A
P ow er Sensor HP/Agilent 8482A
P ow er Sensor HP/Agilent 8485A (83620B /22B/24B,
83623B /L, and 83630B /L)
P ow er Sensor HP/Agilent 8487A (83640B /L and 83650B/L)
Lo w Band P o w er Flatness
Disr e gar d if b eyond your synthesizer's c ap ability.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SERVICE
P o w er Flatness

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Cal Freq Range Lo Bnd (asterisk on)
AtnStep to Cal
band in to the p ow er meter. Refer to the p ow er meter man ual for instructions.
Zero and calibrate the p ow er meter then connect the p ow er sensor to the syn thesizer RF
output.
On the syn thesizer, press
Pwr Mtr ALC Cal .
The syn thesizer measures p ow er atness and calculates a new atness array for the
frequency band and attenuator setting.
On the syn thesizer, press
SAVE and con tinue to the next step.
High Band P o w er Flatness
band in to the p ow er meter. Refer to the p ow er meter man ual for instructions.
Zero and calibrate the p ow er meter then connect the p ow er sensor to the syn thesizer RF
output.
Cal Freq Range Hi Band (asterisk on)
Millimeter Band P o w er Flatness
Disr e gar d if b eyond your synthesizer's c ap ability.
band in to the p ow er meter. Refer to the p ow er meter man ual for instructions.
Zero and calibrate the p ow er meter then connect the p ow er sensor to the syn thesizer RF
output.
Cal Freq Range mm Band (asterisk on)
Related P erformance T ests
P ow er Flatness
In Case of Difficulty
loaded and selected.
No con troller is allo w ed on the bus.
Swept CW Gener ator T r oublesho oting Guide .
P o w er Flatness

No test equipmen t is required for this pro cedure. This pro cedure calibrates the mo dulation
generator assembly b y adjusting the AM and FM DA Cs. The in ternal DVM measures the
DA C output v oltages. The DA C o set is then zero ed with this adjustmen t pro cedure and the
DA C gain is automatically calibrated with calibration constants.
. Let the syn thesizer w arm up for at least one
hour.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SERVICE
NNNNNNNNNNNNNNNNNNNN
AM DAC adjustmen t is automatically selected (asterisk on).

Figure 3-27. A8R111 and A8R105 Adjustment Locations
FM DAC (asterisk on).
Done .
The syn thesizer will display the following message:
store the calibration constants as protected data
(in EEPR OM).
AM/FM D A C Offset and Gain

Related P erformance T ests
In Case of Difficulty
Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series
Swept CW Gener ator T r oublesho oting Guide .
AM/FM D A C Offset and Gain
AM Accuracy · FM Accuracy · 3-74 Adjustments

This pro cedure consists of three adjustmen ts. Calibration constant #277, AM O set, is
adjusted for no c hange in RF output p ow er when AM is enabled but no v oltage is applied.
EXP AM CAL is adjusted for a 10 dB c hange in p ow er when 1.00 V is applied and AM
10 dB/V is selected. LIN AM CAL is adjusted so that a 0.80 V AM input (100%/V AM
selected) causes the RF output to decrease to 4.00% of its initial v alue.
hour.

Figure 3-28. AM Accuracy Adjustment Setup
Digital V oltmeter HP/Agielnt 3456A
F unction Generator HP/Agilent 3325A
P ow er Meter HP/Agilent 436A/7A/8A
P ow er Sensor HP/Agilent 8485A (83620B /22B/24B,
83623B /L, and 83630B /L)
P ow er Sensor HP/Agilent 8487A (83640B /L and 83650B/L)

output is connected to the syn thesizer's AM input.)
POWER LEVEL
AM On/Off 100%/V (asterisk on)
or, for syn thesizers with Option 002, instead press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MOD
Press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SERVICE
ENTER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Modify Cal Use the rotary knob to adjust the calibration constant for a 0 dB reading on
the p ow er meter.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MOD
AM On/Off 100%/V (asterisk o )
or, for syn thesizers with Option 002, instead press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AM On/Off Ext (asterisk o )
AM On/Off 10dB/V (asterisk on)
or, for syn thesizers with Option 002, instead press:
AM On/Off Ext (asterisk on)
AM Type 10dB/V (asterisk on)
the lo cation of A10R35.


Figure 3-29. A10R35 and A10R39 Locations
AM On/Off 10dB/V (asterisk o )
or, for syn thesizers with Option 002, instead press:
AM On/Off Ext (asterisk o )
F or p ow er meters without a relativ e linear mo de, adjust the syn thesizer p ow er lev el for a
3.16 mW p ow er meter reading.
F or p ow er meters with relativ e linear mo de (%), set the p ow er meter to relativ e.
MOD
AM On/Off 100%/V (asterisk on)
or, for syn thesizers with Option 002, instead press:
AM On/Off Ext (asterisk on)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AM Type 100%/V (asterisk on)
for the lo cation of A10R39.
Related P erformance T ests
In Case of Difficulty
Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series
Swept CW Gener ator T r oublesho oting Guide .

This adjustmen t applies to L mo dels with the pre xes listed b elow. This pro cedure uses
calibration constant n um b er 277 to adjust the maximum p ow er out in scalar mo de to match
the RF output amplitude with scalar o .
Mo del Pre x
83623L 3722A and ab ov e
83630L 3722A and ab ov e
83640L 3722A and ab ov e
83650L 3722A and ab ov e
Oscilloscop e HP/Agilent 1740 or HP/Agilent 54600B
Crystal Detector HP/Agilent 3330D (Negative Detector)
hour.
CW
NNNNNNNNNNNNNNNNNNNN
Scalar (no * indicates Scalar is o )
.5V/div or less
adjust o set for trace on screen
NNNNNNNNNNNNNNNNNNNN
Scalar (* indicates Scalar is on)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Adjust Menu . En ter password if requested
(factory password 8360) and then press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Modify Cal using the RPG, arrow k eys, or n umeric k ey pad, c hange the v alue
of cal constant 277 so the maximum p ow er out with scalar on is equal to the pulse p ow er
out with scalar o .
NNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNN
yes .

Figure 3-30. AM Offset Adjustment Setup

In the ALC lo op, the in tegrator input from the ALC detector is dela y ed (phase shifted from
the reference input) through the ALC mo dulator, RF path, and ALC detector. Because of
this, amplitude mo dulation can p eak the AM frequency resp onse. In this adjustmen t, y ou set
the dela y in the reference input equal to the dela y in the input from the ALC detector.
The syn thesizer is set to maximum lev eled p ow er at a CW frequency , with amplitude
mo dulation at the maximum rate. Monitoring the INT L VL signal (the in tegrator output),
adjust the AM DELA Y to minimize an y A C v oltage presen t.

Figure 3-31. AM Delay Adjustment Setup
Digital V oltmeter HP/Agilent 3456A
F unction Generator HP/Agilent 3325A
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Peak RF Always (asterisk on)
Set to 3 dB b elow maximum sp eci ed lev eled p ow er.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MOD
AM On/Off 100%/V (asterisk on)

or, for syn thesizers with Option 002, instead press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AM On/Off Ext (asterisk on)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AM Type 100%/V (asterisk on)
The function generator should ha v e a 50 output for an accurate output.
If necessary , use a BNC tee and monitor the output lev el while the function
generator is connected to the AM input.
Y ou should exp ect to see v ery small c hanges in v oltage.

Figure 3-32. AM Delay Adjustment Location
Related P erformance T ests
In Case of Difficulty
Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series
Swept CW Gener ator T r oublesho oting Guide .

This adjustmen t sets the gain on the FM driver assembly to match the YO FM coil
sensitivit y , and to provide a 10 MHz/V FM input sensitivit y . The function generator is set to
provide FM at a 1 MHz rate and a 2.404 MHz deviation. This corresp onds to a mo dulation
index of 2.404, whic h should result in no p ow er in the carrier. FM gain is adjusted to
minimize the RF output p ow er.
up for at least one hour.

Figure 3-33. FM Gain Adjustment Setup
F unction Generator HP/Agilent 3325A
Sp ectrum Analyzer HP/Agilent 8566B
The function generator sp eci ed is accurate enough to set the amplitude
without using a DVM. If y ou use a di eren t function generator, use a wide
bandwidth DVM, suc h as an HP/Agilent 3458A, to accurately set the v oltage.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MOD
FM On/Off 100kHz (asterisk on)
or, for syn thesizers with Option 002, instead press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FM On/Off Ext (asterisk on)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FM Coupling 100kHz (asterisk on)

Figure 3-34. FM Gain Adjustment Location
Related P erformance T ests
In Case of Difficulty
BNC tee with the function generator connected to the FM input.
Swept CW Gener ator T r oublesho oting Guide .

Use this adjustmen t to set the syn thesizer mo dulation (27.78 kHz square w av e) for equal RF
on/o p erio ds. This is accomplished b y minimizing the ev en harmonics of the 27.78 kHz
square w av e.
w arm up for at least one hour.
Figure 3-35. Square W a v e Symmetry Adjustment Setup
Sp ectrum Analyzer HP/Agilent 8566B
POWER LEVEL
Make sure the signal is cen tered on the display .
MOD
Pulse On/Off Scalar (asterisk on)
or, for syn thesizers with Option 002, instead press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Pulse On/Off Scalar (asterisk on)
(56 kHz aw a y).
display).
Square W a v e Symmetry

sp ectrum analyzer. Press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SERVICE
Use the rotary knob or arrow k eys to select calibration constant #330. Press
and use the rotary knob to adjust the calibration constant.
Lo w Band Adjustment
Disr e gar d if b eyond your synthesizer's c ap ability.
#329 (Slow Sym Lo w Band) for a minimum reading on the sp ectrum analyzer.
to store the calibration constants as protected data (in EEPR OM).
Related P erformance T ests
In Case of Difficulty
Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series
Swept CW Gener ator T r oublesho oting Guide .
Square W a v e Symmetry

No test equipmen t is required for this pro cedure.
This pro cedure sets the AM input imp edance to either 50 or 2 k . The AM input
imp edance on A10 m ust b e set to 2 k for prop er op eration of the in ternal mo dulation
generator (Option 002).
Instrumen ts without Option 002 should p erform step A, \A10 Jump er".
Instrumen ts with Option 002 should set the A10 jump er to the 2 k p osition using step A,
\A10 Jump er" and should p erform step B, \A8 Jump er".

A. A10 Jumper
the jump er to the p osition marked \2K". The jump er can b e returned to the \50 OHM"
p osition to c hange the AM input imp edance to 50 ohms.

Figure 3-36. A10 Jumper Locations
Agilent 8360 Adjustments · AM Input Impedance

B. A8 Jumper
Figure 3-37.
mov e the jump er to the p osition marked \2K". The jump er can b e returned to the p osition
marked \50" to c hange the AM input imp edance to 50 ohms.

Figure 3-37. A8 Jumper Locations
Related P erformance T ests
In Case of Difficulty
If the jump er is missing, see \Replaceable P arts".

A soldering iron is required for this pro cedure.
This pro cedure sets the FM input imp edance to either 50 or 600 . The FM input
imp edance on A11 m ust b e set to 600 for prop er op eration of the in ternal mo dulation
generator (Option 002).
Instrumen ts without Option 002 should p erform step A, \A11 Jump er".
Instrumen ts with Option 002 should set the A11 jump er to the 600 p osition using step A,
\A11 Jump er" and should p erform step B, \A8 Jump er".

A. A11 Jumper
to the 50 ohm p osition. T o c hange it to 600 ohms, remov e the resistor from the A11
assembly b y unsoldering it. The FM input imp edance can b e reset to 50 ohms b y soldering
the resistor bac k in to the A11 assembly .
If y ou will nev er use the 50 ohm setting, the resistor can b e clipp ed out.

Figure 3-38. FM Input Impedance Jumper Locations
3-90 Adjustments · FM Input Impedance

B. A8 Jumper
Figure 3-39 .
mov e the jump er to the p osition marked \600". The jump er can b e returned to the
p osition marked \50" to c hange the FM input imp edance to 50 ohms.

Figure 3-39. A8 Jumper Locations
Related P erformance T ests
Maxim um FM Deviation
In Case of Difficulty
If the jump er is missing, see \Replaceable P arts".

The in ternal mo dulation generator is set up for FM op eration. A 100 kHz rate, measured at
the AM/FM OUTPUT connector, is used as the reference amplitude. A calibration constant
is adjusted so the AM/FM output signal is the same amplitude as for a 1 MHz FM rate.
up for at least one hour.

Figure 3-40. Modulation Generator Flatness Adjustment Setup
Digitizing Oscilloscop e HP/Agilent 54111D
NNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FM On/Off Int (asterisk on)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Internal FM Rate
Internal FM Dev
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRIOR
Time/Division 2 s
Trig Src Chan 1

Display Time 0.2 s
the sine w av e. Do not let the sinew av e b e clipp ed. (Use the front panel k eypad b ecause the
RPG c hanges the sensitivity in steps to o large for this purp ose.)
Signal clipped message is display ed, rep eat the v ertical sensitivit y adjustmen t un til the
message is no longer display ed when the signal is measured.)
Record the amplitude of the sinew av e V p-p.
Internal FM Rate
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ENTER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Modify Cal and, using the rotary knob, c hange the calibration constant un til the
sinew av e display ed on the oscilloscop e is the same amplitude (V p-p) as the recorded v alue.
(The oscilloscop e do es not automatically up date the measuremen t; con tinue to press the
measure function as y ou c hange the calibration constant.)
store the calibration constants as protected data
(in EEPR OM).
Related P erformance T ests
In Case of Difficulty
Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series
Swept CW Gener ator T r oublesho oting Guide .

No test equipmen t is required for this pro cedure.
This pro cedure sets the external trigger to trigger on either the leading or the trailing edge
of a TTL signal. When the jump er (shown in Figure 3-41) is in the POS p osition a sweep is
triggered on the leading edge of the pulse (when the signal c hanges from 0 to 5 v olts). When
the jump er is in the NEG p osition a sweep is triggered on the trailing edge of the pulse (when
the signal c hanges from 5 to 0 v olts).

Figure 3-41. External Trigger Edge Selection Location
Triggering on the Leading Edge
jump er sho wn in Figure 3-41 . In some instrumen ts the jump er p ositions will b e rotated to
. In these instrumen ts the jump er v ertically connects the tw o left p oints in the
POS p osition and the tw o right p oints in the NEG p osition. (POS and NEG p ositions are
lab eled on all assemblies.)
has b een mov ed to the NEG p osition mov e bac k it to the POS p osition. The syn thesizer
no w triggers a sweep on the leading edge
(a 0 to 5 v olt c hange).
Triggering on the Trailing Edge
jump er sho wn in Figure 3-41 . In some instrumen ts the jump er p ositions will b e rotated to
. In these instrumen ts the jump er v ertically connects the tw o left p oints in the
POS p osition and the tw o right p oints in the NEG p osition. (POS and NEG p ositions are
lab eled on all assemblies.)
it mov e the jump er to the NEG p osition. The syn thesizer no w triggers a sweep on the
trailing edge (a 5 to 0 v olt c hange).
External Trigger Edge Selection

No test equipmen t is required for this pro cedure.
A v e-section switch on the A12 m ultiplier/ lter driver assembly sets the v olts/GHz scale
for the rear panel V/GHz output connector. A t the factory , syn thesizers with a maximum
frequency of 30 GHz are set to 0.5 V/GHz and higher frequency instrumen ts are set to 0.25
V/GHz. The following pro cedure describ es ho w to set the switch to scale the output.
follo wing table:
No damage results from setting the v olts/GHz to a scale that is b ey ond the
frequency limits given in the table ho w ev er the v olts/GHz will b e inaccurate
b ey ond the frequency limits.
1 V/GHz 20 GHz closed X
0.5 V/GHz 40 GHz op en op en closed closed X
0.25 V/GHz 50 GHz op en op en op en op en X

Figure 3-42. A12S1 Location
Related P erformance T ests

In Case of Difficulty
Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series
Swept CW Gener ator T r oublesho oting Guide .
3-96 Adjustments · V olts/GHz

No test is required
When replacing the A9 Pulse Board, the E1-E2, E3-E4 jump er needs to b e prop erly
con gured. There is one con guration for all 8360A/B mo dels. There are 2 p ossible
con gurations for 8360L mo dels. The L mo del con guration is dep enden t on the instrumen t
pre x.
On the A9 assembly , set J1 to the con guration for the sp eci c mo del as de ned in the
follo wing table:
Mo del Pre x Jump er
83623L 3722A and ab ov e E3-E4
83630L 3722A and ab ov e E3-E4
83640L 3722A and ab ov e E3-E4
83650L 3722A and ab ov e E3-E4
all other mo dels and pre x E1-E2
A9 Pulse Board Jumper (W1)
Description and Procedure · Agilent 8360 Adjustments

No test is required
After completing the A9 Pulse Board jump er selection, v erify that the following test patc hes
are set for L mo dels with the follo wing pre x and ab ov e. No patc hes are required for L mo dels
b elow the pre x listed
Mo del Pre x
83623L 3722A and ab ov e
83630L 3722A and ab ov e
83640L 3722A and ab ov e
83650L 3722A and ab ov e
107 2 3740 1740
107 5 150 7500
108 1 3740 1740
108 2 150 7500
108 3 6000 3000
108 4 3740 1740
T o Access Selftest P atc h Men u
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
En ter password if requested(factory password is 8360) and then press
NNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

T o Add a T est P atc h
Press:
Specs and the following messages will app ear on the display:
TEST NUMBER TO PATCH:0 En ter the test n um b er (107), then press
TEST DATUM POINT TO PATCH:0 En ter the datum n um b er (2), then press
MIN SPECIFICATION:0 En ter minimum v alue ( 3740), then press
MAX SPECIFICATION:0 En ter maxim um v alue ( 1740), then press
T o Skip a Selftest
Press:
DataSkip the follo wing message will b e display ed:
TEST NUMBER TO PATCH:0 En ter the test n um b er (108), then press
TEST DATUM POINT TO PATCH:0 En ter the datum n um b er (3), then press
T o Delete a T est P atc h
Scroll to the test patc h to delete and press
Delete Patch .



This section con tains the follo wing information on calibration constants and ho w to use them:
Chec ksum V eri cation
Calibration Constant P assword
Changing W orking Data Calibration Constants
Sa ving W orking Data Calibration Constants
Loading Protected Data Calibration Constants
Loading Default Data Calibration Constants
Calibration constants are data whic h describ e y our individual instrumen t. Calibration
constants con tain serial n um b er pre x, hardw are con guration, and b oard revision
information. Calibration constants also con tain calibration information sp eci c to the
op eration of eac h instrumen t that is used to make sure that the instrumen t meets
sp eci cations.
Changing calibration constants can cause y our instrumen t to not meet
sp eci cations.
Agilent 8360 Calibration Constants 4-1

The syn thesizer has three memory areas reserv ed for calibration constants:
W orking data is the set of calibration constants accessed during normal op eration and
con tains the calibration information required for optimum instrumen t p erformance. W orking
data is stored in RAM, and is maintained b y a 1 farad capacitor.
Protected data resides in EEPR OM. This calibration data is essen tially the same as w orking
data, but is not dep enden t on the capacitor. If the syn thesizer's c hec ksum test fails, or if the
capacitor b ecomes disc harged and w orking data is lost, the syn thesizer copies protected data
v alues in to w orking data RAM.
If the calibration constants are defaulted, the syn thesizer will need to b e
completely adjusted b efore it will meet its sp eci cation.
Default data resides in UVEPR OM. This data di ers from w orking and protected data in
that it cannot b e c hanged. This data represen ts a typical syn thesizer, and is not optimized
for y our instrumen t. The default calibration constants are a starting p oint for calibration.
The syn thesizer will probably not meet sp eci cations. The syn thesizer uses default data if
a problem exists in b oth w orking and protected data or if y ou select
Default Cal in the
calibration utility men u.
A t instrumen t p ow er on, the calibration constants stored in RAM are added and compared
with the c hec ksum v alue. If the tw o v alues match, the curren t calibration constants are
maintained in RAM.
If the RAM c hec ksum do es not v erify , the calibration constants are loaded in to RAM from
EEPR OM. These calibration constants are added and compared with their c hec ksum v alue. If
the tw o v alues match, these calibration constants (from EEPR OM) are maintained in RAM.
If the RAM c hec ksum still do es not v erify , the default calibration constants are loaded
in to RAM from UVEPR OM. A message is display ed on the syn thesizer indicating that the
default calibration constants are in use. The syn thesizer will probably not meet p erformance
sp eci cations. All the p erformance tests should b e run.
4-2 Calibration Constants Agilent 8360

Calibration Constant P assw ord
The syn thesizer is shipp ed with a factory-set password. A password disables access to the
adjustmen t men u unless the password is en tered (see \En tering a P assword"). Calibration
constants c annot b e man ually altered without accessing the adjustmen t men u.
The follo wing is the factory-set password: 8360 Y ou can set a new password. See \Setting a
P assword".
Y ou can eliminate the password. See \Disabling a P assw ord".
If y ou ha v e forgotten the password, or if y ou require access to the calibration constant
adjustmen ts for calibration purp oses and do not kno w the password, see \Bypassing the
P assword".
Entering a P assw ord
An asterisk on the
Disable Adjust softkey in the service men u indicates that a password is
set. In order to access the adjustmen t men u y ou m ust en ter that password:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Adjust Menu .
The follo wing message is display ed: ENTER PASSWORD: 0
.
The asterisk on the
Disable Adjust softkey turns o .
Adjust Menu again to access the adjustmen t men u.
En tering the password allo ws p ermanen t access to the adjustmen t men u, ev en if line p ow er is
cycled. T o disable access again, set another password.
Disabling a P assw ord
If y ou wish to eliminate a password (either the factory-set password or one y ou set y ourself ),
p erform the follo wing:
Disable Adjust .
The follo wing message is display ed: ENTER PASSWORD: 0
.
The asterisk on the
Disable Adjust softkey turns o .
Disabling the password allo ws p ermanen t access to the adjustmen t men u, ev en if line p ow er is
cycled.
Agilent 8360 Calibration Constants 4-3

Setting a P assw ord
If a password is already set on the syn thesizer and y ou wish to c hange it, rst follo w the
\Disabling a P assword" pro cedure. This eliminates the curren t password. Then con tinue with
this pro cedure.
If no password is set on the syn thesizer (there is no asterisk on the
Disable Adjust softk ey),
and y ou wish to set a password, p erform the following:
Disable Adjust .
The following message is display ed: SET PASSWORD: 0
T erminate the en try b y pressing:
.
accessible. T o v erify that y ou do wish to set the password, select
Yes, Confirm .
The asterisk on the
Disable Adjust softkey turns on, indicating the password is set.
4-4 Calibration Constants Agilent 8360
·

Bypassing the P assw ord
If y ou require access to the adjustmen t men u for calibration purp oses, a password is set, and
y ou do not kno w the password, p erform the follo wing:
.
The asterisk on the
Disable Adjust softkey turns o enabling access to the adjustmen t
men u.
When access to the adjustmen t men u is no longer desired, enable the previously set password.
P erform the follo wing:
.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
The asterisk on the
Disable Adjust softkey turns on indicating that y ou cannot access
the adjustmen t men u.
If y ou ha v e forgotten the password and wish to set a new one, p erform steps 1 through 3 and
then steps 7 through 10.
On the syn thesizer, select
Disable Adjust .
The follo wing message is display ed: SET PASSWORD: 0
allow ed. T erminate the en try b y pressing:
.
b e accessible. T o v erify that y ou do wish to set the password, select
Yes, Confirm .
Complete steps 4 through 6 to close the switch and enable the password.
Agilent 8360 Calibration Constants 4-5

Changing W orking Data Calibration Constants
If y ou need to mo dify the w orking data calibration constants, the follo wing pro cedure accesses
the calibration constants and lets y ou c hange them.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Calib Menu .
Select Cal .
En ter the n um b er of the calibration constant y ou wish to c hange using the up/do wn
arrow k eys, the rotary knob, or the n umeric k eypad. T erminate n umeric k eypad en tries b y
.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Modify Cal .
Change the v alue of the calibration constant using the right/left arrow k eys, the rotary
knob, or the n umeric k eypad. T erminate n umeric k eypad en tries b y pressing
.
k ey or b y pressing
.
Sa ving W orking Data Calibration Constants
In some adjustmen t pro cedures y ou will c hange w orking data calibration constants. The
follo wing pro cedure stores the calibration constants as protected data (in EEPR OM).
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Cal Util Menu .
NNNNNNNNNNNNNNNNNNNNNNNNNN
Save Cal . A w arning is display ed informing y ou that c hanging the calibration
constants may drastically a ect instrumen t p erformance. T o v erify that y ou do wish to
sav e the new calibration constants, select
yes .
The w orking data calibration constants are no w stored as protected data
(in EEPR OM).
k ey or pressing
.
4-6 Calibration Constants Agilent 8360

Loading Protected Data Calibration Constants
If y our w orking data calibration constants ha v e b een altered or deleted, the follo wing
pro cedure loads the protected calibration constants from EEPR OM in to w orking data
memory .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Cal Util Menu .
Recall Cal .
A w arning is display ed informing y ou that c hanging the calibration constants may
drastically a ect instrumen t p erformance. T o v erify that y ou do wish to recall the
protected data calibration constants, select
yes .
The protected data calibration constants are no w loaded in to w orking data.
k ey or pressing
.
PRESET
Loading Default Data Calibration Constants
If y ou w an t to use the generic syn thesizer calibration constants as w orking data, the following
pro cedure loads the default calibration constants from UVEPR OM in to w orking data memory .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SERVICE
Cal Util Menu .
Default Cal .
A w arning is display ed informing y ou that c hanging the calibration constants may
drastically a ect instrumen t p erformance. T o v erify that y ou do wish to recall the default
yes .
The default data calibration constants are no w loaded in to w orking data.
k ey or pressing
.
Agilent 8360 Calibration Constants 4-7

T able 4-9 provides the following information for eac h calibration constant:
The calibration constant n um b er in ascending order.
The calibration constant name. An asterisk indicates that this calibration constant is
adjusted b y an automated adjustmen t.
The range of realistic adjustmen t (adjustment outside of the range may b e p ossible but
will probably cause the syn thesizer to b e sev erely misadjusted) or the default v alue (some
calibration constants are set to a default v alue and are not adjustable).
The related adjustmen t pro cedure, man ual, automated, or front panel, that adjusts that
calibration constant.
Blank calibration constants are reserv ed for future use.
4-8 Calibration Constants Agilent 8360
Calibration Constant Descriptions

T able 4-9. Calibration Constant Descriptions
Num b er Description Range or
1 Identi er Resp onse 0 | HP/Agilen t 83620A
1 | HP/Agilen t 83621A
2 | HP/Agilen t 83622A
3 | HP/Agilen t 83623A
4 | HP/Agilen t 83624A
5 | HP/Agilen t 83640A
6 | HP/Agilen t 83642A
7 | HP/Agilen t 8340B
8 | HP/Agilen t 83631A
9 | HP/Agilen t 8341B
10 | HP/Agilen t 83650A
11 | HP/Agilen t 83651A
12 | HP/Agilen t 83652A
13 | Agilent 83630A
14 | HP/Agilen t 83632A
15 | Agilent 83620B
16 | Agilent 83621B
17 | Agilent 83622B
18 | Agilent 83623B
19 | Agilent 83624B
20 | Agilent 83640B
21 | Agilent 83631B
22 | Agilent 83650B
23 | Agilent 83651B
24 | Agilent 83630B
25 | Agilent 83623L
26 | Agilent 83630L
27 | Agilent 83640L
28 | Agilent 83650L
2 Dwell After RF On 2
3 User Con guration 0
4 Lo ck out F eatures 0
5 Display Field Lo ck 0
6 Retrace Dwell Time 0
16 Pro duction Info 1 0
17 Pro duction Info 2 0
18 Pro duction Info 3 Instrument sp eci c
Agilent 8360 Calibration Constants 4-9
Related · Default · Adjustment

T able 4-9. Calibration Constant Descriptions (continued)
Num b er Description Range or
19 Pro duction Info 4 Instrument sp eci c
20 Pro duction Info 5 Instrument sp eci c
21 Lo ck Dwell Time 2
22 Max Ph Lk W ait 100
23 YO Slew Rate 500
24 YO Kick Max 6000
25 YO Settle Rate 20
36 YO Dly T erm A Bnd 0 400 to 800 YO Delay Adjustment
37 YO Dly T erm A Bnd 1 400 to 800 YO Delay Adjustment
38 YO Dly T erm A Bnd 2 400 to 800 YO Delay Adjustment
39 YO Dly T erm A Bnd 3 400 to 800 YO Delay Adjustment
40 YO Dly T erm A Bnd 4 400 to 800 YO Delay Adjustment
41 YO Dly T erm A Bnd 5 400 to 800 YO Delay Adjustment
42 YO Dly T erm A Bnd 6 400 to 800 YO Delay Adjustment
43 YO Dly T erm A Bnd 7 400 to 800 YO Delay Adjustment
50 YO Dly T erm B Bnd 0 300 to 700 YO Delay Adjustment
51 YO Dly T erm B Bnd 1 300 to 700 YO Delay Adjustment
52 YO Dly T erm B Bnd 2 300 to 700 YO Delay Adjustment
4-10 Calibration Constants Agilent 8360
Related · Default · Adjustment

T able 4-9. Calibration Constant Descriptions (continued)
Num b er Description Range or
53 YO Dly T erm B Bnd 3 300 to 700 YO Delay Adjustment
54 YO Dly T erm B Bnd 4 300 to 700 YO Delay Adjustment
55 YO Dly T erm B Bnd 5 300 to 700 YO Delay Adjustment
56 YO Dly T erm B Bnd 6 300 to 700 YO Delay Adjustment
57 YO Dly T erm B Bnd 7 300 to 700 YO Delay Adjustment
72 *CW O set 500 to +500 YO Driver Gain And Linearit y
73 YO O V er; Band 0 0
74 YO O V er; Band 1 0
75 YO O V er; Band 2 0
76 YO O V er; Band 3 0
77 YO O V er; Band 4 0
78 YO O V er; Band 5 0
79 YO O V er; Band 6 0
80 YO O V er; Band 7 0
Agilent 8360 Calibration Constants 4-11
Related · Default · Adjustment

T able 4-9. Calibration Constant Descriptions (continued)
Num b er Description Range or
87 YO Swp V er; Band 0 0
88 YO Swp V er; Band 1 0
89 YO Swp V er; Band 2 0
90 YO Swp V er; Band 3 0
91 YO Swp V er; Band 4 0
92 YO Swp V er; Band 5 0
93 YO Swp V er; Band 6 0
94 YO Swp V er; Band 7 0
101 YTM T emp Co ecient 0
105 SRD Bias A; Band 1 255 Amp/Multiplier Adjustments
106 SRD Bias A; Band 2 30 to 230 Amp/Multiplier Adjustments
107 SRD Bias A; Band 3 30 to 230 Amp/Multiplier Adjustments
108 SRD Bias A; Band 4 30 to 230 Amp/Multiplier Adjustments
109 SRD Bias A; Band 5 30 to 230 Amp/Multiplier Adjustments
110 SRD Bias A; Band 6 30 to 230 Amp/Multiplier Adjustments
111 SRD Bias A; Band 7 30 to 230 Amp/Multiplier Adjustments
4-12 Calibration Constants Agilent 8360
Related · Default · Adjustment

T able 4-9. Calibration Constant Descriptions (continued)
Num b er Description Range or
118 SRD Bias B; Band 1 255 Amp/Multiplier Adjustments
119 SRD Bias B; Band 2 30 to 230 Amp/Multiplier Adjustments
120 SRD Bias B; Band 3 30 to 230 Amp/Multiplier Adjustments
121 SRD Bias B; Band 4 30 to 230 Amp/Multiplier Adjustments
122 SRD Bias B; Band 5 30 to 230 Amp/Multiplier Adjustments
123 SRD Bias B; Band 6 30 to 230 Amp/Multiplier Adjustments
124 SRD Bias B; Band 7 30 to 230 Amp/Multiplier Adjustments
131 Squegg Clamp 1A 15 to 200 Ampli er/Filter Adjustments
132 Squegg Clamp 2A 255 Ampli er/Filter Adjustments
133 Squegg Clamp 3A 255 Ampli er/Filter Adjustments
134 Squegg Clamp 4A 255 Ampli er/Filter Adjustments
135 Squegg Clamp 5A 255 Ampli er/Filter Adjustments
136 Squegg Clamp 6A 255 Ampli er/Filter Adjustments
137 Squegg Clamp 7A 255 Ampli er/Filter Adjustments
144 YTM Dly T erm A Hrm 1 600 to 2000 Ampli er/Filter Adjustments
145 YTM Dly T erm A Hrm 2 600 to 2500 Ampli er/Filter Adjustments
146 YTM Dly T erm A Hrm 3 600 to 3000 Ampli er/Filter Adjustments
147 YTM Dly T erm A Hrm 4 600 to 2500 Ampli er/Filter Adjustments
148 YTM Dly T erm A Hrm 5 600 to 3000 Ampli er/Filter Adjustments
149 YTM Dly T erm A Hrm 6 600 to 3000 Ampli er/Filter Adjustments
150 YTM Dly T erm A Hrm 7 600 to 3000 Ampli er/Filter Adjustments
Agilent 8360 Calibration Constants 4-13
Related · Default · Adjustment

T able 4-9. Calibration Constant Descriptions (continued)
Num b er Description Range or
157 YTM Dly T erm B Hrm 1 200 to 500 Ampli er/Filter Adjustments
158 YTM Dly T erm B Hrm 2 200 to 500 Ampli er/Filter Adjustments
159 YTM Dly T erm B Hrm 3 200 to 500 Ampli er/Filter Adjustments
160 YTM Dly T erm B Hrm 4 200 to 500 Ampli er/Filter Adjustments
161 YTM Dly T erm B Hrm 5 200 to 500 Ampli er/Filter Adjustments
162 YTM Dly T erm B Hrm 6 200 to 500 Ampli er/Filter Adjustments
163 YTM Dly T erm B Hrm 7 200 to 500 Ampli er/Filter Adjustments
170 YTM Bx Dly T erm A 1 600 to 2000 Amp/Multiplier Adjustments
171 YTM Bx Dly T erm A 2 600 to 2500 Amp/Multiplier Adjustments
172 YTM Bx Dly T erm A 3 600 to 3000 Amp/Multiplier Adjustments
173 YTM Bx Dly T erm A 4 600 to 2500 Amp/Multiplier Adjustments
174 YTM Bx Dly T erm A 5 600 to 3000 Amp/Multiplier Adjustments
175 YTM Bx Dly T erm A 6 600 to 3000 Amp/Multiplier Adjustments
176 YTM Bx Dly T erm A 7 600 to 3000 Amp/Multiplier Adjustments
4-14 Calibration Constants Agilent 8360
Related · Default · Adjustment

T able 4-9. Calibration Constant Descriptions (continued)
Num b er Description Range or
183 YTM Bx Dly T erm B 1 200 to 500 Amp/Multiplier Adjustments
184 YTM Bx Dly T erm B 2 200 to 500 Amp/Multiplier Adjustments
185 YTM Bx Dly T erm B 3 200 to 500 Amp/Multiplier Adjustments
186 YTM Bx Dly T erm B 4 200 to 500 Amp/Multiplier Adjustments
187 YTM Bx Dly T erm B 5 200 to 500 Amp/Multiplier Adjustments
188 YTM Bx Dly T erm B 6 200 to 500 Amp/Multiplier Adjustments
189 YTM Bx Dly T erm B 7 200 to 500 Amp/Multiplier Adjustments
195 YTM Kick Threshold 110 Amp/Multiplier Adjustments
196 YTM CW Kick Max 600 Amp/Multiplier Adjustments
197 YTM Mono Band Kick 2000 Amp/Multiplier Adjustments
198 YTM Stereo Band Kick 2000 Amp/Multiplier Adjustments
199 YTM Slew Rate 400 Amp/Multiplier Adjustments
200 YTM Slew Max 50
201 YTM Slew Min 0
202 YTM Neg Kick W ait 2
203 YTM Fwd Kick Pct 50
204 YTM Fwd Kick W ait 30
205 YTM Rise; Band 1 1 to 20 Ampli er/Filter Adjustments
206 YTM Rise; Band 2 A1 20 to 25 Ampli er/Filter Adjustments
207 YTM Rise; Band 2 B1 1 to 4 Ampli er/Filter Adjustments
208 YTM Rise; Band 2 A2 10 to 20 Ampli er/Filter Adjustments
209 YTM Rise; Band 3 A1 5 to 255 Ampli er/Filter Adjustments
210 YTM Rise; Band 3 B1 0 to 50 Ampli er/Filter Adjustments
211 YTM Rise; Band 3 A2 5 to 255 Ampli er/Filter Adjustments
212 YTM Rise; Band 3 B2 0 to 50 Ampli er/Filter Adjustments
213 YTM Rise; Band 3 A3 1 to 20 Ampli er/Filter Adjustments
Agilent 8360 Calibration Constants 4-15
Related · Default · Adjustment

T able 4-9. Calibration Constant Descriptions (continued)
Num b er Description Range or
214 YTM Rise; Band 4 1 to 50 Ampli er/Filter Adjustments
215 YTM Rise; Band 5 1 to 50 Ampli er/Filter Adjustments
216 YTM Rise; Band 6 1 to 50 Ampli er/Filter Adjustments
217 YTM Rise; Band 7 1 to 50 Ampli er/Filter Adjustments
225 *YTM Gain; Band 1 1500 to 2500 Amp/Multiplier Adjustments;
Auto T rac k
226 *YTM Gain; Band 2 1000 to 2500 Amp/Multiplier Adjustments;
Auto T rac k
227 *YTM Gain; Band 3 500 to 2500 Amp/Multiplier Adjustments;
Auto T rac k
228 *YTM Gain; Band 4 500 to 2500 Amp/Multiplier Adjustments;
Auto T rac k
229 *YTM Gain; Band 5 500 to 2500 Amp/Multiplier Adjustments;
Auto T rac k
230 *YTM Gain; Band 6 500 to 2500 Amp/Multiplier Adjustments;
Auto T rac k
231 *YTM Gain; Band 7 500 to 2500 Amp/Multiplier Adjustments;
Auto T rac k
238 *YTM O set; Band 1 1500 to 2500 Amp/Multiplier Adjustments;
Auto T rac k
239 *YTM O set; Band 2 1500 to 3000 Amp/Multiplier Adjustments;
Auto T rac k
4-16 Calibration Constants Agilent 8360
Related · Default · Adjustment

T able 4-9. Calibration Constant Descriptions (continued)
Num b er Description Range or
240 *YTM O set; Band 3 1500 to 3000 Amp/Multiplier Adjustments;
Auto T rac k
241 *YTM O set; Band 4 1500 to 3000 Amp/Multiplier Adjustments;
Auto T rac k
242 *YTM O set; Band 5 1500 to 3000 Amp/Multiplier Adjustments;
Auto T rac k
243 *YTM O set; Band 6 1500 to 3000 Amp/Multiplier Adjustments;
Auto T rac k
244 *YTM O set; Band 7 1500 to 3000 Amp/Multiplier Adjustments;
Auto T rac k
248 YTM B2 O set O set 50 to +50 Amp/Multiplier Adjustments
249 Autotrac k D A C Setting 2048
251 *L VL D A C Gain Hi Bnd 100 to +100 ALC P o wer Level
252 *L VL D A C Gain Lo Bnd 100 to +100 ALC P o wer Level
253 *L VL D A C Gain Xtal 100 to +100
254 *L VL D A C Gain 2ndOut 0
255 *L VL D A C Gain PwrMtr 100 to +100
256 *L VL D A C Gain Mo dule 100 to +100
257 *L VL D A C Gain ALC O 0
264 *L VL D A C Ofs Hi Bnd 50 to +300 ALC P o wer Level
Agilent 8360 Calibration Constants 4-17

T able 4-9. Calibration Constant Descriptions (continued)
Num b er Description Range or
265 *L VL D A C Ofs Lo Bnd 100 to 250 ALC P o wer Level
266 *L VL D A C Ofs Xtal 1000 to +1000
267 *L VL D A C Ofs 2ndOut 0
268 *L VL D A C Ofs PwrMtr 100 to +100
269 *L VL D A C Ofs Mo dule 100 to +100
270 *L VL D A C Ofs ALC O 0
274 In ternal FM Flat 10,000 to +10,000 Mo dulation Generator
275 FM D A C Gain T rim 1000 to +1000
276 AM D A C Gain T rim 1000 to +1000
277 AM O set 20 to +20
277 AM O set (L Mo dels) 620 to +20 AM O set (L Mo dels)
(serial pre xes 3722
283 *ALC Det Ofs; Hi 100 to 160 ALC P o wer Level
284 *ALC Det Ofs; Lo 100 to 160 ALC P o wer Level
285 *ALC Det Ofs; Xtal 100 to 160
286 *ALC Det Ofs; 2nd Out 100 to 160
292 *ALC Log Brkpt; Hi 0 to 255 ALC P o wer Level
4-18 Calibration Constants Agilent 8360

T able 4-9. Calibration Constant Descriptions (continued)
Num b er Description Range or
293 *ALC Log Brkpt; Lo 0 to 255 ALC P o wer Level
294 *ALC Log Brkpt; Xtal 0 to 255
295 *ALC Log Brkpt; 2nd 0 to 255
301 *ALC Mo d Gain; Band 0 100 to 200 Mo dulator O set And Gain
302 *ALC Mo d Gain; Band 1 100 to 200 Mo dulator O set And Gain
303 *ALC Mo d Gain; Band 2 70 to 200 Mo dulator O set And Gain
304 *ALC Mo d Gain; Band 3 70 to 200 Mo dulator O set And Gain
305 *ALC Mo d Gain; Band 4 100 to 200 Mo dulator O set And Gain
306 *ALC Mo d Gain; Band 5 50 to 150 Mo dulator O set And Gain
307 *ALC Mo d Gain; Band 6 50 to 150 Mo dulator O set And Gain
308 *ALC Mo d Gain; Band 7 50 to 150 Mo dulator O set And Gain
315 ALC Mo d Ofs; Band 0 50 to 150 Mo dulator O set And Gain
316 ALC Mo d Ofs; Band 1 50 to 150 Mo dulator O set And Gain
317 ALC Mo d Ofs; Band 2 150 to 220 Mo dulator O set And Gain
318 ALC Mo d Ofs; Band 3 150 to 220 Mo dulator O set And Gain
319 ALC Mo d Ofs; Band 4 100 to 200 Mo dulator O set And Gain
320 ALC Mo d Ofs; Band 5 100 to 200
321 ALC Mo d Ofs; Band 6 100 to 200
322 ALC Mo d Ofs; Band 7 100 to 200
Agilent 8360 Calibration Constants 4-19

T able 4-9. Calibration Constant Descriptions (continued)
Num b er Description Range or
329 Slo w Sym Lo w Band 100 to 160
330 Slo w Sym High Band 100 to 160
331 A ttenuator Sw P oin t 1000
335 Squegg Clamp 1B 100
336 Squegg Clamp 2B 100
337 Squegg Clamp 3B 100
338 Squegg Clamp 4B 100
339 Squegg Clamp 5B 100
340 Squegg Clamp 6B 100
341 Squegg Clamp 7B 100
351 ADC 5 V Range +OFF 5000 to +5000 ADC Adjustment
352 ADC 15 V Range +OFF 5000 to +5000 ADC Adjustment
353 ADC 5 V Range +GAIN 1950 to +1950 ADC Adjustment
354 ADC 15 V Range +GAIN 750 to +750 ADC Adjustment
4-20 Calibration Constants Agilent 8360
Related · Default · Adjustment

T able 4-9. Calibration Constant Descriptions (continued)
Num b er Description Range or
355 ADC 5 V Range OFF 5000 to +5000 ADC Adjustment
356 ADC 15 V Range OFF 5000 to +5000 ADC Adjustment
357 ADC 5 V Range GAIN 750 to +750 ADC Adjustment
358 ADC 15 V Range GAIN 1750 to +1950 ADC Adjustment
367 PwrMtr Gain; Lo Band 0
368 PwrMtr Gain; Hi Band 0
369 Pwr Mtr O set 0
373 YTM Dly T erm C Hrm 1 0
374 YTM Dly T erm C Hrm 2 0
375 YTM Dly T erm C Hrm 3 0
376 YTM Dly T erm C Hrm 4 0
377 YTM Dly T erm C Hrm 5 0
378 YTM Dly T erm C Hrm 6 0
379 YTM Dly T erm C Hrm 7 0
382 YTM Bx Dly T erm C 1 0
383 YTM Bx Dly T erm C 2 0
384 YTM Bx Dly T erm C 3 0
385 YTM Bx Dly T erm C 4 0
386 YTM Bx Dly T erm C 5 0
Agilent 8360 Calibration Constants 4-21
Related · Default · Adjustment

T able 4-9. Calibration Constant Descriptions (continued)
Num b er Description Range or
387 YTM Bx Dly T erm C 6 0
388 YTM Bx Dly T erm C 7 0
4-22 Calibration Constants Agilent 8360
Related · Default · Adjustment

T able 4-9. Calibration Constant Descriptions (continued)
Num b er Description Range or
440 Doubler Amp Mo de 0
441 Ph uo c Magic Enable 0
442 *Sweeptime Range 0 3750 to 6250 Sweep Ramp
443 *Sweeptime Range 1 19655 to 32755 Sweep Ramp
444 *Sweeptime Range 2 9360 to 15600 Sweep Ramp
449 D Amp Clamp 1 75 to 250 Ampli er/Filter Adjustments
450 D Amp Clamp 2 75 to 250 Ampli er/Filter Adjustments
Agilent 8360 Calibration Constants 4-23
Related · Default · Adjustment

T able 4-9. Calibration Constant Descriptions (continued)
Num b er Description Range or
451 D Amp Clamp 3 75 to 250 Ampli er/Filter Adjustments
452 D Amp Clamp 4 75 to 250 Ampli er/Filter Adjustments
453 D Amp Clamp 5 75 to 255
454 D Amp Clamp 6 75 to 255
455 D Amp Clamp 7 75 to 255
461 Serial Num b Pre x See fo otnote 1
463 Hardw are Con g 1 See fo otnote 1
464 Hardw are Con g 2 See fo otnote 1
473 P o wer Sp ec Band 0 See fo otnote 2
474 P o wer Sp ec Band 1 See fo otnote 2
475 P o wer Sp ec Band 2 See fo otnote 2
476 P o wer Sp ec Band 3 See fo otnote 2
477 P o wer Sp ec Band 4 See fo otnote 2
478 P o wer Sp ec Band 5 See fo otnote 2
479 P o wer Sp ec Band 6 See fo otnote 2
480 P o wer Sp ec Band 7 See fo otnote 2
4-24 Calibration Constants Agilent 8360

T able 4-9. Calibration Constant Descriptions (continued)
Num b er Description Range or
487 Max Sweep Rate 300
488 Max Sweep Rate 8516 100
493 A1 Revision See fo otnote 3
494 A2 Revision See fo otnote 3
495 A3 Revision See fo otnote 3
496 A4 Revision See fo otnote 3
497 A5 Revision See fo otnote 3
498 A6 Revision See fo otnote 3
499 A7 Revision See fo otnote 3
500 A8 Revision See fo otnote 3
501 A9 Revision See fo otnote 3
502 A10 Revision See fo otnote 3
503 A11 Revision See fo otnote 3
504 A12 Revision See fo otnote 3
505 A13 Revision See fo otnote 3
506 A14 Revision See fo otnote 3
507 A15 Revision See fo otnote 3
508 A16 Revision See fo otnote 3
509 A17 Revision See fo otnote 3
510 A18 Revision See fo otnote 3
511 A19 Revision See fo otnote 3
512 A20 Revision See fo otnote 3
513 A21 Revision See fo otnote 3
514 A22 Revision See fo otnote 3
515 A23 Revision See fo otnote 3
Agilent 8360 Calibration Constants 4-25
Related · Default · Adjustment

T able 4-9. Calibration Constant Descriptions (continued)
Num b er Description Range or
525 Hib ernation Time 90
526 Bandcross Time 1 42
527 Bandcross Time 2 32
528 Bandcross Time 3 28
529 Bandcross Time 4 26
530 Bandcross Time 5 28
531 Bandcross Time 6 20
532 Bandcross Time 7 20
536 Sp ecial Option 1 0
537 Sp ecial Option 2 0
538 Sp ecial Option 3 0
539 Sp ecial Option 4 0
540 Sp ecial Option 5 0
541 Sp ecial Option 6 0
542 Sp ecial Option 7 0
543 Sp ecial Option 8 0
544 Sp ecial Option 9 0
545 Sp ecial Option 10 0
4-26 Calibration Constants Agilent 8360
Related · Default · Adjustment

T able 4-9. Calibration Constant Descriptions (continued)
Num b er Description Range or
554 Display FIFO Time 245
555 Display FSCC Time 1
561 YTF Dly T erm A Hrm 1 1000
562 YTF Dly T erm A Hrm 2 1000
563 YTF Dly T erm A Hrm 3 1000
564 YTF Dly T erm A Hrm 4 1000
565 YTF Dly T erm A Hrm 5 1000
566 YTF Dly T erm A Hrm 6 1000
567 YTF Dly T erm A Hrm 7 1000
571 YTF Dly T erm B Hrm 1 300
572 YTF Dly T erm B Hrm 2 300
573 YTF Dly T erm B Hrm 3 300
574 YTF Dly T erm B Hrm 4 300
575 YTF Dly T erm B Hrm 5 300
576 YTF Dly T erm B Hrm 6 300
577 YTF Dly T erm B Hrm 7 300
Agilent 8360 Calibration Constants 4-27
Related · Default · Adjustment

T able 4-9. Calibration Constant Descriptions (continued)
Num b er Description Range or
581 YTF Dly T erm C Hrm 1 0
582 YTF Dly T erm C Hrm 2 0
583 YTF Dly T erm C Hrm 3 0
584 YTF Dly T erm C Hrm 4 0
585 YTF Dly T erm C Hrm 5 0
586 YTF Dly T erm C Hrm 6 0
587 YTF Dly T erm C Hrm 7 0
591 YTF Bx Dly T erm A 1 1000
592 YTF Bx Dly T erm A 2 1000
593 YTF Bx Dly T erm A 3 600 to 2500
594 YTF Bx Dly T erm A 4 600 to 2500
595 YTF Bx Dly T erm A 5 1000
596 YTF Bx Dly T erm A 6 1000
597 YTF Bx Dly T erm A 7 1000
601 YTF Bx Dly T erm B 1 300 Ampli er/Filter Adjustments
602 YTF Bx Dly T erm B 2 300 Ampli er/Filter Adjustments
603 YTF Bx Dly T erm B 3 200 to 450 Ampli er/Filter Adjustments
604 YTF Bx Dly T erm B 4 200 to 450 Ampli er/Filter Adjustments
605 YTF Bx Dly T erm B 5 300 Ampli er/Filter Adjustments
606 YTF Bx Dly T erm B 6 300 Ampli er/Filter Adjustments
607 YTF Bx Dly T erm B 7 300 Ampli er/Filter Adjustments
4-28 Calibration Constants Agilent 8360
Related · Default · Adjustment

T able 4-9. Calibration Constant Descriptions (continued)
Num b er Description Range or
611 YTF Bx Dly T erm C 1 0
612 YTF Bx Dly T erm C 2 0
613 YTF Bx Dly T erm C 3 0
614 YTF Bx Dly T erm C 4 0
615 YTF Bx Dly T erm C 5 0
616 YTF Bx Dly T erm C 6 0
617 YTF Bx Dly T erm C 7 0
621 *YTF Gain; Band 1 1500 to 2500 Ampli er/Filter Adjustments
622 *YTF Gain; Band 2 1000 to 2500 Ampli er/Filter Adjustments
623 *YTF Gain; Band 3 500 to 2500 Ampli er/Filter Adjustments
624 *YTF Gain; Band 4 500 to 2500 Ampli er/Filter Adjustments
625 *YTF Gain; Band 5 500 to 2500 Ampli er/Filter Adjustments
626 *YTF Gain; Band 6 500 to 2500 Ampli er/Filter Adjustments
627 *YTF Gain; Band 7 500 to 2500 Ampli er/Filter Adjustments
631 *YTF O set; Band 1 1500 to 2500 Ampli er/Filter Adjustments
632 *YTF O set; Band 2 1500 to 3000 Ampli er/Filter Adjustments
633 *YTF O set; Band 3 1500 to 3000 Ampli er/Filter Adjustments
634 *YTF O set; Band 4 1500 to 3000 Ampli er/Filter Adjustments
635 *YTF O set; Band 5 1500 to 3000 Ampli er/Filter Adjustments
636 *YTF O set; Band 6 1500 to 3000 Ampli er/Filter Adjustments
637 *YTF O set; Band 7 1500 to 3000 Ampli er/Filter Adjustments
Agilent 8360 Calibration Constants 4-29
Related · Default · Adjustment

T able 4-9. Calibration Constant Descriptions (continued)
Num b er Description Range or
641 A24 Rev;Low Band See fo otnote 3
642 A25 Rev;Mo d Splitter See fo otnote 3
643 A26 Rev;YIG Osc See fo otnote 3
644 A27 Rev;Isolator See fo otnote 3
645 A28 Rev;SYTM See fo otnote 3
646 A29 Rev;Amp/Detector See fo otnote 3
647 A30 Rev;Coupler See fo otnote 3
648 A31 Rev;Attenuator See fo otnote 3
649 A32 Rev;Doubler See fo otnote 3
650 A33 Rev;Ampli er See fo otnote 3
651 A34 Rev;RP Assem bly See fo otnote 3
652 A35 Rev;Isolator See fo otnote 3
653 A36 PM/A42 PMSF Rev See fo otnote 3
655 A38 Rev;Dual Mo d See fo otnote 3
660 Quick Step Slew 0 120 | A13 Revision=4
5 | All others
661 Quick Step Slew 1 120 | A13 Revision=4
5 | All others
662 Quick Step Slew 2 60 | A13 Revision=4
10 | All others
663 Quick Step Slew 3 40 | A13 Revision=4
15 | All others
664 Quick Step Slew 4 30 | A13 Revision=4
17 | All others
665 Quick Step Slew 5 20 | A13 Revision=4
26 | All others
666 Quick Step Slew 6 20 | A13 Revision=4
26 | All others
667 Quick Step Slew 7 15 | A13 Revision=4
35 | All others
4-30 Calibration Constants Agilent 8360
Related · Default · Adjustment

T able 4-9. Calibration Constant Descriptions (continued)
Num b er Description Range or
670 RF In terface ID 0
671 A9 Pulse Board ID See fo otnote 3
672 A12 YTM Driver ID See fo otnote 3
673 A28 Amp/Mult ID See fo otnote 3
674 A29 Amp/Filt ID See fo otnote 3
675 A38 Dual Mo d ID See fo otnote 3
676 A36 PM/A42 PMSF ID See fo otnote 3
677 A32 Doubler ID See fo otnote 3
678 A10 ALC Board ID See fo otnote 3
Sp eci c to the syn thesizer's hardw are con guration. The calibration constant v alue is set at
the factory . See ***: undefined*** .
Sp eci ed maximum lev eled p ow er for the band (for example, 7 dBm equals 700, 10 dBm
equals 1000).
Sp eci c to the syn thesizer's hardw are con guration. The calibration constant ID n um b er
and revision v ary dep ending on the assembly v ersion. See T able 4-11 .
Agilent 8360 Calibration Constants 4-31
Related · Default · Adjustment

T able 4-10. Hardw are Configuration Cal Constants #463 and #464

83620B +19,600 128 +4 +4096 +2 { +1 +2
83622B +19,584 128 +4 +4096 +2 { +1 +2
83623B +19,608 128 +4 +4096 6 127 +1 +2
83624B +19,592 128 +4 +4096 6 { +1 +2
83630B +19,600 128 +4 +4096 +2 { +1 +2
83640B +19,632 128 +4 +4096 +2 127 +1 +2
83650B +19,888 128 +4 +4096 +2 { +1 +2
83621B +30,864 {
83631B +30,864 {
83651B +32,176 {
83623L +27,792 128 { +4096 { { +1 {
83630L +27,792 128 { +4096 { { +1 {
83640L +27,824 128 { +4096 { { +1 {
83650L +28,080 128 { +4096 { { +1 {
1 Option H10 and all other options are m utually exclusiv e.
19,600 - 128 + 4096 = 23,568
4-32 Calibration Constants Agilent 8360

T able 4-11.
Assemblies and Corresponding Cal Constant ID Numbers and Revisions

Assembly ID Numb er Revision Part Numb er Description
Agilent 8360 Calibration Constants 4-33

T able 4-11.
Assemblies and Corresponding Cal Constant ID Numbers and Revisions

Assembly ID Numb er Revision Part Numb er Description
1 08360-60264 CPU (Battery) 2 R OM Chip Set
4-34 Calibration Constants Agilent 8360
(continued)

T able 4-11.
Assemblies and Corresponding Cal Constant ID Numbers and Revisions
Assembly ID Numb er Revision Part Numb er Description
1 0 5086-7596 Switched Amp/Mult ( 83621B/31B)
2 1 5086-7622 Amp/Mult ( 83630B/L
2 2 5086-7622 Amp/Mult with Aux Out ( 83620B,
A29 #674 #646
1 1 5086-7623 Amp/Filter
2 1 5086-7636 Amp/Filter
( 83630B/40B/50B Option 006)
3 0 5086-7616 Amp/Filter ( 83620B/22B/23B/24B)
4 1 5086-7637 Amp/Filter ( 83623L)
0 0955-0148 Directional Coupler
( 83623B/L and 83624B)
0 5086-7645 Bridge Detector
( 83620B/22B and 83630B/L)
0 0955-0125 Directional Coupler ( 83631B )
0 5086-7656 Bridge Detector
( 83640B/L and 83650B/L)
0 0955-0523 Directional Coupler ( 83651B )
0 33326-60003 Step A tten uator ( 83620B/22B/24B,
0 33326-60002 Step A tten uator ( 83640B/L)
0 33326-60005 Step A tten uator ( 83650B/L)
Agilent 8360 Calibration Constants 4-35

T able 4-11.
Assemblies and Corresponding Cal Constant ID Numbers and Revisions
Assembly ID Numb er Revision Part Numb er Description
0 1 5086-7607 F requency Doubler ( 83640B /L)
1 1 5086-7604 F requency Doubler ( 83650B /L)
1 1 5086-7513 F requency Doubler ( 83651B )
0 5086-7465 Ampli er (High P ow er) ( 83623B )
0 5086-7510 Ampli er (High P ow er) ( 83624B )
0 08360-60061 RP Floating BNC (Option 004)
0 08360-60186 RP Floating BNC (Option 002 and 004)
A37
A38 #675 #655
1 0 5086-7602 Dual Mo dulator ( 83620B,
83624B, 83630B/L,
2 0 5086-7634 AM Mo dulator
(Option 006 and 83621B/31B/51B)
2 0 5087-7017 Mo d Doubler Amp
4-36 Calibration Constants Agilent 8360

T able 4-11.
Assemblies and Corresponding Cal Constant ID Numbers and Revisions
Assembly ID Numb er Revision Part Numb er Description
0 0 ( 83620B/22B/24B
1 1 ( 83630B/L, 83640B/L,
0 1 5086-7628 Pulse Mo d Switched Filter
( 83620B/22B Option 006)
0 1 5086-7629 Pulse Mo d Switched Filter
( 83623B/24B Option 006)
1 1 5086-7630 Pulse Mo d Switched Filter
( 83630B Option 006)
1 1 5086-7631 Pulse Mo d Switched Filter
( 83640B/50B Option 006)
Agilent 8360 Calibration Constants 4-37
(continued) · and 83623B /L) · and 83650B /L)



This c hapter explains ho w to load and run the \Agilent 8360 Service Support Softw are"
revision A.02.00. These tests require op erator in teraction.
This softw are requires that the syn thesizer is set to the SCPI programming language. This
is set b y the rear panel switch or b y the front panel. Refer to the \Installation" c hapter of
y our syn thesizer's user's guide for further information. After using the softw are, return the
syn thesizer to its original setting.
This c hapter con tains the follo wing information on using the softw are:
Setting Up the System
Installing the Softw are
Running the Softw are
Automated P erformance T ests
Step A tten uator Flatness T est
P ow er Flatness and Accuracy T est
Step A tten uator Flatness Adjustmen t
YO Delay Adjustmen t
P ow er Flatness Adjustmen t
F ront P anel Em ulation
P ow er Sensor Con guration and Calibration F actor File
Agilent 8360 Automated T ests 5-1

Setting Up the System
The automated tests require an HP 9000 series 200/300 desktop computer with at least 2.25
megabytes of RAM, a disk drive, and an GPIB in terface. This program will not run with
high resolution color monitors ( > 512 x 390 pixels). Softw are is provided on tw o 3.5 inc h
disks, formatted double-sided, and are usable in double-sided disk drives only . An y required
measuremen t instrumen ts are listed in eac h test pro cedure.
The test softw are requires BASIC v ersion 5.1 and the following binaries.
T able 5-1. Required Binaries
Language Extensions Driv ers
CLOCK CR T A or CR TB (dep ends on CR T)
COMPLEX CS80 or DISC (dep ends on disk)
EDIT HPIB
ERR SRM
GRAPH
GRAPHX
IO
KBD
MA T
MS
PDEV
TRANS
XREF
Refer to the BASIC user's do cumen tation for instructions on loading the op erating system
and binaries.
5-2 Automated T ests Agilent 8360
Hardw are Requirements · Operating System Requirements

Installing the Softw are
The automated tests are provided on tw o double-sided disks. Previous revisions may b e
di eren t.
Mak e W orking Copies
Before doing an ything else, make a w orking cop y of the master disks!
The master disk is shipp ed from the factory write-protected and cannot b e written to or
initialized in this mo de. W e recommend y ou maintain this master disk in write-protect mo de.
During execution of some tests, the program reads from and writes to the disk, use a w orking
cop y that is not write-protected when y ou run the softw are.
Installing the Program onto a Hard Disk
W e recommend that y ou run the softw are from a hard disk. It is p ossible to op erate it from a
single or dual opp y disk drive, ho w ev er sp eed will b e sacri ced.
T o install the program on to y our hard disk, create a directory that will con tain the program
and its asso ciated les. (Refer to the BASIC user's do cumen tation for instructions on creating
directories.) After y ou create the directory , run the install program b elow to load all of the
les on to y our hard disk.
disk. In this example, the source lo cation is :,700,0 and the destination lo cation is :,1400,0
Y our hard disk is no w set up to run the automated tests.
Operating from a Flopp y Disk
When y ou run the softw are from a opp y disk drive, b e sure to leav e the second master disk
in the disk drive while the program runs. If this is not done, the program will not run.
Agilent 8360 Automated T ests 5-3
Disk Files

Running the Softw are
This softw are uses RAM memory v olumes for fast access of data les. These
v olumes may also b e used b y other programs and could con tain data that will
b e erased b y this program. Make sure that the computer y ou are using do es
not ha v e imp ortant data in an y memory v olumes b efore running this softw are.
The computer con taining the softw are m ust b e the only con troller on the bus. If more than
one con troller is presen t, the softw are will not run prop erly .
Loading the Softw are
copied to a w ork disk (if running from a opp y disk).
Use the BASIC MSI command. See the BASIC L anguage R efer enc e for more information
on setting the default mass storage.
CAPS LOCK ON or OFF cannot b e c hanged while the program is running.
Set the k eyb oard b efore y ou load the softw are.
Insert master disk #1 in y our default drive, or from y our hard-driv e directory , p erform the
.
to start the test program. If y ou are using opp y disks, the program will
prompt y ou to install the second disk.
av ailable mo del n um b ers. Using the arrow k eys, select the correct mo del to b e tested.
.
En ter the appropriate information.
test. Press
.
Refer to the follo wing pages for a detailed description and sp eci c op erating
information (if an y) for a sp eci c test.
test information. When the test nishes, y ou are returned to the test men u.
Either select another test to run or select the
DONE softkey to exit the program.
5-4 Automated T ests Agilent 8360
Configuration Limitations · Press · Press

P erformance v eri cation test for Option 001
A p ow er meter is used to make relative p ow er measuremen ts to determine the actual
attenuation of eac h attenuator card. Mismatch error is minimized b y including a xed
attenuation in front of the card b eing measured for b oth parts of the relativ e measuremen t
(with/without the card). The softw are uses the v alue of eac h attenuator card to calculate
what the actual attenuation is when set to a sp eci c v alue. F or example, the attenuator v alue
for the 30 dB setting is the sum of the 10 dB and the 20 dB cards. The e ect of digital ALC
correction is accoun ted for b y the softw are.
The follo wing test equipmen t is required for this p erformance test. No substitutions are
allo w ed.
T able 5-2. T est Equipment Required
10 dB A tten uator 8493C
30 dB A tten uator 11708A
P ow er Meter 438A
P ow er Sensor (50 MHz to 26.5 GHz) 8485A
P ow er Sensor (50 MHz to 26.5 GHz) 8485D
P ow er Sensor (100 kHz to 4.2 GHz) 8482A
1 Substitute HP/Agilent 8487A/D for 40 GHz synthesizers.
Agilent 8360 Automated T ests 5-5

Connect the equipmen t as sho wn in Figure 5-1 . Preset all instrumen ts and let them w arm up
for at least one hour.
This program requires that p ow er sensor calibration factors b e stored in a le.
If the calibration factors ha v e not b een en tered previously , refer to the utility ,
\4. P ow er Sensor Con guration and Calibration F actor File".
Select the step attenuator atness test and follow the prompts on the display .

Figure 5-1. Step A ttenuator Flatness T est Setup
5-6 Automated T ests Agilent 8360
Procedure

P erformance v eri cation test for standard instrumen t
If Option 001 is installed in y our instrumen t, p erform the \Step A tten uator
Flatness" test.
This test uses the user atness correction array to measure p ow er atness at the RF output.
The p ow er is measured ev ery 10 MHz in low band and ev ery 100 MHz in high and millimeter
bands.
The calibration factors for the p ow er sensor m ust b e added to the program b efore the test
is run. Refer to the utility , \4. P ow er Sensor Con guration and Calibration F actor File", if
necessary .
The follo wing test equipmen t is required for this test. No substitutions are allo w ed. (The test
setup is given in the softw are.)
T est Equipment Required
P ow er Meter 438A
P ow er Sensor (100 kHz to 4.2 GHz) 8482A
P ow er Sensor (50 MHz to 26.5 GHz) 8485A
P ow er Sensor (50 MHz to 50 GHz)
10 dB A tten uator (DC to 26.5 GHz)
1 F or 40 GHz synthesizers only .
2 F or high p ow er synthesizers only .
Select the p ow er atness and accuracy test and en ter y our p ow er sensor data b y p erforming
the follo wing steps.
If the p ow er sensor data w as added previously , and is still curren t, skip the
follo wing steps and run the test.
HELP softkey to access the p ow er sensor con guration and calibration men us.
Agilent 8360 Automated T ests 5-7

Do not edit the \Sensor ID". The names set at the factory m ust remain
unc hanged for the program to run prop erly . The factory recommended
\Zero Hr" is 1.00 and the \Cal Hr" is 24.00. Use the left and right arrows to
mov e the cursor within a eld. Use the up and do wn arrows to incremen t or
decremen t the v alue.
RETURN selects the eld.
the program.
en ter the data.
5-8 Automated T ests Agilent 8360

store the calibration factors for future use.
Agilent 8360 Automated T ests 5-9

Calibration for Option 001
Before doing an y adjustmen t, en ter the password for access to the adjustmen t calibration
constants. Use the front panel or the front panel em ulation program provided with this
softw are.
See Chapter 4 for information on enabling the calibration constants with the password.
A p ow er meter is used to make relative p ow er measuremen ts to determine the actual
attenuation of eac h attenuator card. Mismatch error is minimized b y including a xed
attenuation in front of the card b eing measured for b oth parts of the relative measuremen t
(with/without the card). The softw are uses the v alue of eac h attenuator card to calculate
what the actual attenuation is when set to a sp eci c v alue. F or example, the attenuator v alue
for the 30 dB setting is the sum of the 10 dB and the 20 dB cards. The e ect of digital ALC
correction is accoun ted for b y the softw are.
The following test equipmen t is required for this adjustmen t. No substitutions are allo w ed.
T able 5-3. T est Equipment Required
10 dB A tten uator 8493C
30 dB A tten uator 11708A
P ow er Meter 438A
P ow er Sensor (50 MHz to 26.5 GHz) 8485A
P ow er Sensor (50 MHz to 26.5 GHz) 8485D
P ow er Sensor (100 kHz to 4.2 GHz) 8482A
1 Substitute HP/Agilent 8487A/D for 40 GHz synthesizers.
Connect the equipmen t as sho wn in Figure 5-2 . Preset all instrumen ts and let them w arm up
for at least one hour.
Select the step attenuator atness adjustmen t and follow the prompts on the display .
5-10 Automated T ests Agilent 8360

Figure 5-2. Step A ttenuator Flatness Adjustment Setup
Agilent 8360 Automated T ests 5-11

This adjustmen t uses the in ternal coun ter to measure swept frequency accuracy . YO dela y
a ects the swept frequency accuracy at fast sweep times (less than 100 ms). This pro cedure
adjusts the calibration constants for the YIG oscillator to ac hiev e consistent swept frequency
accuracy ov er the en tire op erating range of the instrumen t.
If y our instrumen t has Option 006, b e certain that y ou iden ti ed y our option
con guration when the softw are w as initially loaded or else this pro cedure will
not run prop erly .
Select the YO dela y adjustmen t and follo w the prompts on the display .
No test equipmen t is required.
5-12 Automated T ests Agilent 8360
Description

This adjustmen t uses an external digital v oltmeter (DVM) to measure V COMP on the YO
driver assembly and compares the reading to the ADC measuremen t. The ADC calibration
constants are adjusted so that the DVM and ADC measuremen ts are the same. The
pro cedure is rep eated for eac h v oltage range (b oth plus and minus supplies).
Select the ADC adjustmen t and follo w the prompts on the display .
The follo wing test equipmen t is required for this adjustmen t. (The test setup is given in the
softw are.)
T est Equipment Required
Instrumen t HP/Agilent Mo del Numb er
Digital V oltmeter 3456A, 3457A, or 3458A
Agilent 8360 Automated T ests 5-13
Description

Calibration for standard instrumen t
If Option 001 is installed in y our instrumen t, p erform the \Step A tten uator
Flatness" adjustmen t.
This adjustmen t zeros the digital ALC array and measures the p ow er from the RF output.
The p ow er is measured ev ery 10 MHz in low band and ev ery 100 MHz in high and millimeter
bands. The calibration constants are adjusted to ac hiev e a at output.
The calibration factors for the p ow er sensor m ust b e added to the program. Refer to the
utilit y , \4. P ow er Sensor Con guration and Calibration F actor File", if necessary .
The following test equipmen t is required for this adjustmen t. No substitutions are allo w ed.
(The test setup is given in the softw are.)
T est Equipment Required
P ow er Meter 438A
P ow er Sensor (100 kHz to 4.2 GHz) 8482A
P ow er Sensor (50 MHz to 26.5 GHz) 8485A
P ow er Sensor (50 MHz to 50 GHz)
10 dB A tten uator (DC to 26.5 GHz)
1 F or 40 GHz synthesizers only .
2 F or high p ow er synthesizers only .
Select the p ow er atness adjustmen t and en ter y our p ow er sensor data b y p erforming the
follo wing steps.
If the p ow er sensor data w as added previously , and is still curren t, skip the
following steps and run the test.
HELP softkey to access the p ow er sensor con guration and calibration men us.
5-14 Automated T ests Agilent 8360

Do not edit the \Sensor ID". The names set at the factory m ust remain
unc hanged for the program to run prop erly . The factory recommended
\Zero Hr" is 1.00 and the \Cal Hr" is 24.00. Use the left and right arrows to
mov e the cursor within a eld and the up and do wn arrows to incremen t or
decremen t the v alue.
selects the eld.
the program.
en ter the data.
Agilent 8360 Automated T ests 5-15

store the calibration factors for future use.
5-16 Automated T ests Agilent 8360

This utility simulates an HP/Agilent 8360 syn thesizer front-panel k eyb oard in an instrumen t.
T o access the front panel em ulation utility , select the front panel em ulation program in the
test men u.
The front panel em ulation program cannot run front panel functions that use a p ow er meter
or prin ter. These functions cause the Agilent 8360 B- or L-series syn thesizer to act as a
con troller, whic h con icts with the computer as the con troller on the bus. Use a substitute
front panel to run these functions, or use the tests provided with this softw are.
When using the front panel em ulation program, en ter the letter directly ab ov e the caret ( ^) to
activ ate the function or k ey . Note that the k eys are case sensitive (\L" means \ALC" and \l"
means \LOCAL"). See T able 5-4 for a detailed listing of hardk ey designations.
F or n umeric en try , tw o di eren t terminator k eys are com bined on one hardk ey . The actual
terminator dep ends on the active mo de of the syn thesizer.
en ters 10 ms. When frequency
span is the active en try ,
en ters 10 kHz.
Softkeys are designated b y menukey X, where X is 1 through 5 and corresp onds to the v e
softkey p ositions b elow the display . T o activ ate the softkey , press the function k ey on the
k eyb oard that corresp onds to the n um b er X indicated b y men uk ey .
Agilent 8360 Automated T ests 5-17

T able 5-4. Synthesizer Hardk ey Emulation
Hardkey Program Displa y T o Activ ate
mo d m
ALC
MARKER
USER CAL
SER VICE
PRIOR
SYSTEM
SA VE
saV e V
MENU
USER DEFINED
MENU
ASSIGN
ENTR Y
ENTR Y ON/OFF
entry o f
5-18 Automated T ests Agilent 8360

T able 5-4. Synthesizer Hardk ey Emulation (continued)
Hardkey Program Displa y T o Activ ate
SWEEP TIME
SINGLE
CONT
MENU
mEn u E
CW
ST AR T
STOP
CENTER
SP AN
MENU
POWER
PO WER LEVEL
FL TNESS ON/OFF
RF ON/OFF
MENU
INSTR UMENT ST A TE
PRESET
LOCAL
lo cal l
Agilent 8360 Automated T ests 5-19

The calibration constants utilit y provides a v ariety of w ays to manipulate the calibration
constants stored in the instrumen t.
Use this utilit y to prin t out a list of the calibration constants stored in the instrumen t, to
make a bac k-up of the calibration constants, and to restore calibration constants from a
bac k ed-up le.
En ter the password to allo w access to the calibration constants b y running the front panel
em ulation program provided with this softw are. See Chapter 4 for information on accessing
the calibration constants with the password.
Y ou may need to p erform sev eral steps to reac h y our nal goal. F or example, to mov e the
calibration constants from the instrumen t's w orking memory to a disk for bac k-up, y ou m ust
rst mov e them to the computer memory , and then from computer memory to disk. T o store
the calibration constants, y ou may use the hard disk drive, Disk 2 of y our w orking disks, or a
separate data disk. When storing data on a new disk, b e sure to format the disk b efore y ou
b egin the utility .
Select the calibration constants en try in the test men u and follo w the prompts.
This utility is executed using the front panel em ulation program. A full self-test of the
syn thesizer will b e executed and the results display ed on the computer.
Occasionally , the self-tests may fail when it is run remotely or when external
cables are connected to the instrumen t. If the self-tests fail, remov e all cables
and run the self-tests from the front panel. If the self-tests pass under these
conditions, the instrumen t is op erating prop erly .
External connections to the instrumen t during self-test may not allo w the lines
to c hange conditions within the limits of the self-test's parameters.
5-20 Automated T ests Agilent 8360
Description · Procedure · Description

This utility is used for creating and editing the les that store p ow er sensor calibration
factors. The p ow er sensor calibration factors are used in the \Pow er Flatness" and \Step
A tten uator" p erformance tests and adjustmen t pro cedures. This data can b e stored to the
directory or disk for future use. The p ow er meter/p ow er sensor con guration utilit y is used
to de ne whic h p ow er sensors, b y serial n um b er, are actually used when the automated test
is run. Although y ou may en ter and store calibration factors for sev eral p ow er sensors, the
program will use only the p ow er sensors that y ou ha v e iden ti ed b y serial n um b er in the
con guration utilit y .
Loading Sensor Calibration F actors
The service softw are program for loading sensor calibration factors is unfriendly . F ollowing the
steps outlined should result in a successful en try .
Do not h urry or skip steps.
T o Load Sensor Cal F actors:
\Step A tten uator Flatness T est", \Pow er Flatness Adjust", or \Step A tten uator Flatness
Adjust".
Men u" where the \Pmtr Con g" (Pow er Meter Con guration) and \Pmtr Cal F actors"
selections are lo cated.
only used to select sensor serial n um b ers (mo del n um b ers are xed). After selecting \Pmtr
Con g" the screen will display a list of mo del and serial n um b ers the program will use
during the test.
Imp ortan t Notes for Editing the Selections:
No new mo del n um b ers can b e added. The program only recognizes the mo del n um b ers
listed and will stop if other mo del n um b ers are en tered.
Mo del n um b ers can not b e c hanged. Only the serial n um b ers can b e c hanged in this
men u. Use serial n um b ers to track di eren t sensor mo del n um b ers.
Only edit sensor serial n um b ers in this men u.
Always exit the program b y selecting [17] \Sav e Con guration" and, then, pressing
\Done". A bug in the program may erase mo del n um b ers if this is not done. If a mo del
n um b er is missing create a new w orking disk n um b er 2 using the master disk or purge
\PMTRCONF" le from the w orking disk and cop y the \PMTRCONF" le bac k to the
w orking disk from the master disk.
The frequency range of the instrumen t determined whic h sensor mo del n um b ers listed
will b e used during the test. The serial n um b er determines whic h cal factors will b e
used. The follo wing list explains whic h mo del n um b ers are used when.
Agilent 8360 Automated T ests 5-21
Description

F or instrumen ts with a maxim um frequency of 26.5 GHz, the follo wing sensors are used:
8485D + 10 dB pad
8485A
8482A
F or instrumen t with a maximum frequency ab ov e 26.5 GHz, the follo wing sensors are
used:
8487D + 10 dB pad
8487A
8482A
After Selection, Con tinue with Cal Sensor Pro cedures:
Con guration", and press \Select".
added, scroll to the mo del n um b er to b e edited and press \Select" to get to the \Edit
Sensor #1" screen.
selection [4] and press \Select". Once \Select" has b een pressed, a n um b er m ust b e
en tered.
n um b er and press \Return", then \Done". The mo del and new serial n um b er should b e
listed in the table.
\Sav e P ow er Meter Con guration" and pressing \Select" then \Done" to exit the routine.
n um b ers, scroll to \Pmtr Cal F actor" selection [3], and press \Select". A list of serial
n um b ers and sensor mo del n um b ers should b e display ed.
n um b er is not display ed, return to the \Pmtr Con g" program and rep eat en tering the
correct serial n um b er.
When en tering cal factors for a new serial n um b er, the message \F ailed attempt to access
le . . . " will app ear. Press \Ab ort". Read the message display ed and then press \OK".
an explanation of eac h selection.
n um b er selected. After \Select" is pressed the cal factors are display ed. Scroll to the
line to b e edited and press \Select". There are four c hoices for editing the cal factors.
V alue - V alue is used to edit a cal factor. En ter the new v alue b y typing ov er the old
data. Do not use the delete k ey .
F requency - F requency is used to edit the frequency . En ter the new v alue b y typing
ov er the old data. Do not use the delete k ey .
5-22 Automated T ests Agilent 8360

Insert - Inserts a line in the table at the lo cation of the p ointer.
Delete - Deletes the line in the table at the lo cation of the p ointer.
[2] Edit Sensor Information: Not used.
sensor form disk or the directory sp eci ed in the MSI command.
[4] Store Cal F actors: This stores the cal factors for the selected p ow er sensor to disk or
the directory sp eci ed b y the MSI command.
[5] Curren t: Displa ys the curren t sensor serial and mo del n um b er.
selected p ow er sensor.
T o en ter frequency and correction factors for the sensor, en ter the frequency , a
comma, and then the correction factor (F requency , Correction). The frequency m ust
b e en tered in GHz (.01 = 10 MHz). Correction factors m ust b e en tered in % (97.0 =
97.0%). En ter frequencies in ascending order (low est to highest).
The rst frequency p oint m ust b e b elow the start frequency the sensor is used at: F or
rst frequency p oints en ter:
8485A/D
8482A, 8485A/D
There m ust b e a calibration factor at 50 MHz. The test softw are requires this 50
MHz calibration factor for the p ow er sensor.
The last frequency p oint m ust b e greater than the highest frequency the sensor is
used at. The last frequency p oints en tered:
8487A/D 50.0 GHz
8482A >4.2 GHz
8485A/D, 8487A/D >26.5 GHz
T o end serial en try en ter \0,0".
Select \Store Cal F actors".
desired c hoice and select
After all data is en tered and \Done" has b een selected the program should con tinue
running. If a \Time Out" error o ccurs, c hec k the setup and press \Retry". If \Retry"
fails to start the program press \Ab ort" and start the program ov er.
Agilent 8360 Automated T ests 5-23

The following men u c hoices are display ed:
This is the instrumen t calibration factors utility . Use it to store and retrieve instrumen t
calibration factors. The follo wing men u c hoices will allow y ou to en ter and store calibration
factors:
[1] Review CALS in Disk File
[2] Review CALS from UUT W orking CALS
[3] Store UUT W orking CALS to Disk File
[4] Load CALS F rom a Disk File to UUT W orking CALS
[5] Cop y UUT W orking CALS to UUT Protected CALS
[6] Cop y UUT Protected CALS to UUT W orking CALS
Pm tr Con g
This is the p ow er meter con guration utility . Use it to en ter serial n um b ers for the p ow er
sensors whic h are listed in the \Sensor ID" column. F or example, when the program requires
an HP/Agilent 8487D p ow er sensor, it will use the calibration factor data that is stored for
whic hev er p ow er sensor has b een listed in the \Serial" column. If HP/Agilen t 8487A/D p ow er
sensors are used to test syn thesizers with maximum stop frequencies of 26.5 GHz, en ter
their serial n um b ers for the HP/Agilent 8485A/D c hoices. The \Sensor ID" and \Name"
columns iden tify the frequency range and p ow er lev el ov er whic h the p ow er sensor is used.
The HP/Agilen t 8487A/D can substitute for an HP/Agilent 8485D; ho w ev er, the rev erse is
not true.
After y ou select a p ow er sensor, mov e the arrow to \Serial Num b er" and en ter the new serial
n um b er. Then select
DONE .
When the p ow er meter con guration is correct, scroll do wn to selection 17 and sav e the
con guration data.
Pm tr Cal F actors
This is the p ow er meter calibration factor utilit y . Use it to en ter and store calibration factors
for eac h p ow er sensor. Notice that the test softw are requires calibration factors at 50 MHz
and at one frequency higher than the sp eci ed stop frequency of the syn thesizer under test.
T o edit or en ter new calibration factors, select the serial n um b er for the p ow er sensor from the
display ed list. (This list is deriv ed from the p ow er meter con guration utility and the p ow er
sensor m ust already b e en tered in the con guration.) The calibration factors whic h w ere
previously stored will b e loaded and the editing men u will b e display ed.
5-24 Automated T ests Agilent 8360
Menu Ov erview · Calco Utility

Agilent 8360 Menu Maps 6-1
Menu Maps

6-2 Menu Maps Agilent 8360



















Disassembly and Replacement Procedures
This c hapter provides the following disassembly and reassembly pro cedures. Use these
pro cedures while repairing or replacing an assembly .
\F ront P anel Disassembly and Reassem bly"
\Rear P anel Disassembly and Reassem bly"
\RF Dec k Disassembly and Reassem bly"
\Motherb oard Disassem bly and Reassem bly"
Agilent 8360 Disassembly and Replacement Procedures 7-1
Introduction

These servicing instructions are for use b y qualified personnel only . T o a v oid
electrical shock, do not perform any servicing unless y ou are qualified to do
so.
The opening of co v ers or remo v al of parts is lik ely to expose dangerous
v oltages. Disconnect the instrument from all v oltage sources while it is being
opened.
The detachable po w er cord is the instrument disconnecting device. It
disconnects the mains circuits from the mains supply before other parts of
the instrument. The front panel switch is only a standb y switch and is not a
LINE switch.
The po w er cord is connected to internal capacitors that may remain liv e for 5
seconds after disconnecting the plug from its po w er supply .
This is a Safety Class I product (pro vided with a protectiv e earthing ground
incorporated in the po w er cord). The mains plug shall only be inserted in a
sock et outlet pro vided with a protectiv e earth contact. Any interruption of the
protectiv e conductor, inside or outside the instrument, is lik ely to mak e the
instrument dangerous. Intentional interruption is prohibited.
For continued protection against fire hazard replace line fuse only with same
type and rating (F 5A/250V). The use of other fuses or material is prohibited.
P erform the following pro cedures only at a static safe w ork station. The
attenuators and prin ted circuit assemblies in this instrumen t are v ery
sensitive to static electricit y damage. W ear an an ti-static wrist strap that is
connected to earth ground.
When reassembling the instrumen t, the cables sho wn in Figure 7-1 m ust b e
routed through the instrumen t as shown or instrumen t p erformance may b e
a ected.
7-2 Disassembly and Replacement Procedures Agilent 8360


Figure 7-1. Cable Routing Locations
Eac h pro cedure lists the required to ols at the b eginning of that section. See Chapter 9 for the
asso ciated part n um b ers.
Agilent 8360 Disassembly and Replacement Procedures 7-3
T ools Required

Front P anel Disassembly and Reassembly
T-10 TORX Screwdriv er (torque 10 in-lb)
0.050 Hex Driver
9/16 in. Nut Driv er
5/16 in. Op en End W renc h
Coax Extractor T o ol{Submin D Connector
Some steps in this pro cedure may not apply to all option con gurations. If a particular step
do es not apply , con tinue with the next appropriate step.
T o disassemble the front panel, follow this pro cedure in the order given.
T o reassemble the front panel, follo w this pro cedure in the rev erse order.
Refer to the o w c hart in Figure 7-2 to determine the necessary disassembly steps for a given
assembly .
7-4 Disassembly and Replacement Procedures
T ools Required · Description and Procedure · Front P anel


Figure 7-2. Front P anel Disassembly Flo w chart
Agilent 8360 Disassembly and Replacement Procedures
Front P anel

Y ou can easily blow the fuse for the 13.5 V supply on the preregulator
assembly if y ou do not disconnect the A C p ow er cord. Also, the danger of
sho c k is increased if the p ow er cord is connected.
An RF braid is in eac h of the side c hannels on the top of the instrumen t.
When y ou reassemble the instrumen t, b e sure this braid remains in the
c hannel.
screwdriver in either slot in the trim strip and lift to remov e, see Figure 7-3.)

Figure 7-3. Front Edge Trim Strip
7-6 Disassembly and Replacement Procedures
Front P anel

Refer to Figure 7-4.
If y ou are removing the front panel only so that y ou can lift up the RF dec k
and if y ou ha v e a at surface in front of the syn thesizer on whic h to lay the
front panel, let the front panel lie in front of the syn thesizer with all the cables
connected and con tinue with step 1 of \2. Lift Up the RF Dec k". If y ou do
not ha v e a at surface for the front panel, disconnect ribb on cables W2 and
W3 from the motherb oard and lay the front panel on top of the syn thesizer.

Figure 7-4. Front P anel Screw Locations
Agilent 8360 Disassembly and Replacement Procedures
Front P anel

Refer to Figure 7-5.
Instrumen ts with Option 004 do not ha v e all the cables sho wn.
002), disconnect the four coaxial cables (W6, W7, W8, and W9) attached to the BNC
connectors.

Figure 7-5. Front P anel Coaxial and Ribbon Cable Locations
7-8 Disassembly and Replacement Procedures
Front P anel

Refer to Figure 7-6.
cable, and the six-wire display cable W1.

Figure 7-6. Front P anel Processor Cables
Agilent 8360 Disassembly and Replacement Procedures
Front P anel

Refer to Figure 7-7.

Figure 7-7. Display Remo v al
7-10 Disassembly and Replacement Procedures
Front P anel

Refer to Figure 7-8.
Pro cessor Remo v al" b efore con tinuing.

Figure 7-8. Source Module Interface Remo v al
Agilent 8360 Disassembly and Replacement Procedures
Front P anel

Refer to Figure 7-9.
Pro cessor Remo v al" b efore con tinuing.

Figure 7-9. RPG Remo v al
7-12 Disassembly and Replacement Procedures
Front P anel

Refer to Figure 7-10 for front panel output.
T o av oid damaging the connector, do not disassemble the RF output
connector assembly . Only unscrew the connector on the end of the cable.
assembly .

Figure 7-10. Remo ving RF Output Connector (Front P anel)
When y ou re-attach the connector and cable to the syn thesizer, rst, lo osen
the n ut b et w een cable W43 and the connector. Then rev erse steps 1 and 2.
Finally , tighten the n ut b et w een cable W43 and the connector to 10 in-lb.
Agilent 8360 Disassembly and Replacement Procedures
Front P anel

Refer to Figure 7-11.
Remo v al" and \5. A2 Source Mo dule In terface Remo v al" b efore con tinuing.
panel.
If the w asher on the front of the connector is not cen tered prop erly , lo osen the
n ut to slide the connector out.

Figure 7-11. Front P anel BNC Locations
7-14 Disassembly and Replacement Procedures
Front P anel

Refer to Figure 7-12.
Remo v al", \4. A21 Alphan umeric Display Remo v al" and \5. A2 Source Mo dule In terface
Remo v al" b efore con tinuing.
It is imp ortant that the top of the k eyb oard b e k ept clean. Dirt on the
con tacts will a ect the p erformance of the k eypad.

Figure 7-12. Keyboard Assembly Remo v al
Refer to Figure 7-13.
Agilent 8360 Disassembly and Replacement Procedures
Front P anel

The follo wing steps can degrade k eypad p erformance. If y ou separate the
an ti-ro c k sheets from the k eypad, y ou m ust replace b oth the an ti-ro c k sheets
and the k eypad.

Figure 7-13. Keypad and Anti-rock Sheets
7-16 Disassembly and Replacement Procedures
Front P anel

Refer to Figure 7-14.
Remo v al", \4. A21 Alphan umeric Display Remo v al", \5. A2 Source Mo dule In terface
Remo v al" and \9. A1 Keyb oard Remo v al" b efore con tinuing.

Figure 7-14. Display Filter
Agilent 8360 Disassembly and Replacement Procedures
Front P anel

Refer to Figure 7-15.
one inc h, place the frame face do wn.
in to place.
same heigh t.

Figure 7-15. Keyboard Reassembly
Refer to Figure 7-16
7-18 Disassembly and Replacement Procedures
Front P anel


Figure 7-16. Keyboard Assembly Screw Locations
Agilent 8360 Disassembly and Replacement Procedures
Front P anel

F ollo w the o w c hart, Figure 7-2, in rev erse order to p erform the reassembly of the follo wing
assemblies:
J1 RF Output Connector
A3 F ront P anel Pro cessor
A21 Alphan umeric Display
A2 Source Mo dule In terface Assem bly
RPG1 Rotary Pulse Generator (RPG)
7-20 Disassembly and Replacement Procedures
BNC Connectors · Displa y Filter · Front P anel

Rear P anel Disassembly and Reassembly
9/16 in. So c k et
T-10 T orx Screwdriv er
T-15 T orx Screwdriv er
5.5 mm Nut Driver
5/16 in. Op en End W renc h
14 mm Op en End W renc h
7 mm Nut Driver
#1 x 3 1/8 in. P ozidrive Screwdriv er
#2 x 4 in. P ozidriv Screwdriv er
Coax Extractor T o ol{Submin D Connector
Some steps in this pro cedure may not apply to all option con gurations. If a particular step
do es not apply , con tinue with the next appropriate step.
T o disassemble the rear panel and asso ciated assemblies, follo w this pro cedure in the order
given.
T o reassemble the rear panel and asso ciated assemblies, follo w this pro cedure in the rev erse
order.
Cable routing may a ect instrumen t p erformance. Refer to Figure 7-1 for a
diagram of the instrumen t cable routing.
Refer to the o w c hart in Figure 7-17 to determine the necessary disassembly steps for a given
assembly .
Agilent 8360 Disassembly and Replacement Procedures
T ools Required · Description and Procedure · Rear P anel


Figure 7-17. Rear P anel Disassembly Flo w chart
7-22 Disassembly and Replacement Procedures
Rear P anel

Y ou can easily blow the fuse for the 13.5 V supply on the preregulator
assembly if y ou do not disconnect the A C p ow er cord. Also, the danger of
sho c k is increased if the p ow er cord is connected.
An RF braid is in eac h of the side c hannels on the top and b ottom of the
instrumen t. When y ou reassemble the instrumen t, b e sure the braids remain
in the side co v er c hannels.
Agilent 8360 Disassembly and Replacement Procedures
Rear P anel

Refer to Figure 7-18.
brac k ets (item 10).

Figure 7-18. B1 Rear P anel F an Assembly
7-24 Disassembly and Replacement Procedures
Rear P anel

Refer to Figure 7-19.
REF OUTPUT, V OL TS/GHz, and TRIGGER OUTPUT.
OUTPUT, PULSE SYNC OUT, and PULSE VIDEO OUT.

Figure 7-19. Rear P anel BNC Connector Locations
Agilent 8360 Disassembly and Replacement Procedures
Disconnection · Rear P anel

Refer to Figure 7-20.
T o av oid damaging the connector, do not disassemble the RF output
connector assembly . Only unscrew the connector on the end of the cable.
instrumen ts with Option 001).

Figure 7-20. Remo ving RF Output Connector
When y ou re-attach the connector and cable to the syn thesizer, rst, lo osen
the n ut b et w een cable W43 and the connector. Then p erform the disassembly
steps in rev erse order. Finally , torque the n ut b et w een cable W43 and the
connector J1 to 10 in-lb (112 N-cm).
7-26 Disassembly and Replacement Procedures
Rear P anel

Refer to Figure 7-21.
con tinuing.

Figure 7-21. Rear P anel Remo v al
Agilent 8360 Disassembly and Replacement Procedures
Rear P anel

Refer to Figure 7-22.
mo dule in terface.
source mo dule in terface cable from the A19 assembly .

Figure 7-22. Rear P anel Cable Locations
7-28 Disassembly and Replacement Procedures
Rear P anel

Refer to Figure 7-23.
P anel Remo v al" b efore con tinuing.

Figure 7-23. Reference Standard Mountings
Agilent 8360 Disassembly and Replacement Procedures
Rear P anel

Refer to Figure 7-24.
P anel Remo v al" b efore con tinuing.

Figure 7-24. Transformer Remo v al
7-30 Disassembly and Replacement Procedures
Rear P anel

Refer to Figure 7-25.
P anel Remo v al" b efore con tinuing.
ALC assembly . (W23 is not included with replacemen t A19 assemblies.)
extractor to ol and then remov e the tw o screws holding the shield to the rear panel
in terface.

Figure 7-25. A19 Rear P anel Interface Remo v al
Agilent 8360 Disassembly and Replacement Procedures
Rear P anel

Refer to Figure 7-26.
P anel Remo v al" b efore con tinuing.

Figure 7-26. V oltage Selection Switch Remo v al
7-32 Disassembly and Replacement Procedures
Rear P anel

Refer to Figure 7-27.

Figure 7-27. Fuse, Line Filter, and Ground Wires
T o reassemble, add 3/8 in. shrink tubing to eac h wire b efore soldering. Then
p erform the ab ov e steps in rev erse order.
Agilent 8360 Disassembly and Replacement Procedures
Rear P anel

Refer to Figure 7-28.
P anel Remo v al" b efore con tinuing.
the switch (this provides clearance for the wrenc h to remov e the fuse holder).

Figure 7-28. Rear P anel Fuse Housing
T o reassemble, add 3/8 in. shrink tubing to eac h wire b efore soldering. Then
p erform the ab ov e steps in rev erse order.
7-34 Disassembly and Replacement Procedures
Rear P anel

Refer to Figure 7-29.
P anel Remo v al" b efore con tinuing.

Figure 7-29. Line Filter Remo v al
T o reassemble, add 3/8 in. shrink tubing to eac h wire b efore soldering. Then
p erform the ab ov e steps in rev erse order.
Agilent 8360 Disassembly and Replacement Procedures
Rear P anel

Refer to Figure 7-30.
panel.

Figure 7-30. Option 004 BNC Connectors Remo v al
7-36 Disassembly and Replacement Procedures
Rear P anel

RF Deck Disassembly and Reassembly
#1 x 3 1/8 in. P ozidriv Screwdriv er
T-10 T orx Screwdriv er
5/16 in. T orque W renc h
5 mm Ball and So c k et Driver
On the RF dec k, all P ozidriv screws ha v e English threads and all T orx screws
ha v e metric threads.
T o disassemble the RF dec k and asso ciated assemblies, follo w this pro cedure in the order
given.
T o reassemble the RF dec k and asso ciated assemblies, follo w this pro cedure in the rev erse
order.
Refer to the o w c hart in Figure 7-31 to determine the necessary disassembly steps for a given
assembly .
Agilent 8360 Disassembly and Replacement Procedures
T ools Required · Description and Procedure · RF Deck


Figure 7-31. RF Deck Disassembly Flo w chart
7-38 Disassembly and Replacement Procedures
RF Deck

Y ou can easily blow the fuse for the 13.5 V supply on the preregulator
assembly if y ou do not disconnect the A C p ow er cord. Also, the danger of
sho c k is increased if the p ow er cord is connected.
An RF braid is in eac h of the side c hannels on the top of the instrumen t.
When y ou reassemble the instrumen t, b e sure this braid remains in the
c hannel.
If y our instrumen t has front panel RF output, remov e the front panel. (See \F ront P anel
Disassembly and Reassem bly", step 2.)
If y our instrumen t has rear panel RF output (Option 004), see step 4 of \Rear P anel
Disassembly and Reassem bly" then con tinue with step 4 of this pro cedure.
T o av oid damaging the connector, do not disassemble the RF output
connector assembly . Only unscrew the connector on the end of the cable.
Agilent 8360 Disassembly and Replacement Procedures
RF Deck


Figure 7-32. Remo ving RF Output Connector
When y ou re-attach the connector and cable to the syn thesizer, rst lo osen the
n ut b et w een cable W43 and the connector. Then rev erse steps 1 through 3.
Finally , torque the n ut b et w een cable W43 and connector J1 to 10 in-lb (112
N-cm).
w atch out that y ou do not damage the cable cen ter conductor).
(you may need to lo osen the other end of W51 also).
7-40 Disassembly and Replacement Procedures
RF Deck


Figure 7-33. W51 RF Cable Location
Av oid damage to the v arious ribb on cables and their connectors attached to
the RF dec k assembly when removing that assembly in the next step. Mov e
all exible coaxial cables out of the w ay of the RF dec k assembly comp onen ts.
Some exible cables will still b e attached prev en ting the RF dec k from b eing
completely remov ed from the instrumen t.
Refer to Figure 7-34.
The analog extender b oard and coax extender cable men tioned in the follo wing
steps are part of the service to ol kit. Refer to Chapter 9 for the part n um b er
of this kit.
Agilent 8360 Disassembly and Replacement Procedures
RF Deck

the RF dec k. Place the RF dec k on the extender b oard and use the tw o hex screws to
attach the dec k to the syn thesizer frame (service p osition). If y ou are not troublesho oting
and are lifting up the RF dec k for disassembly purp oses, attac h the dec k to the
syn thesizer frame without using the extender b oard.
a. A ttac h a coax extender cable b et w een W51 and the low pass lter FL2.
b. Reconnect the RF connector and W43.
i. Lo osen the n ut b et w een cable W43 and the connector.
ii. Reconnect cable W43 at directional coupler A30 (or the attenuator A31 for
instrumen ts with Option 001).
iii. T orque the n ut b et w een cable W43 and connector J1 to 10 in-lb (112 N-cm).

Figure 7-34. RF Deck in the Service P osition
7-42 Disassembly and Replacement Procedures
RF Deck

Refer to Figure 7-35.
attaching the dec k to the syn thesizer frame in order to remov e the cables on the low er edge
of the RF dec k.)
W36 ( 83640B /L and 83650B /L only)
W38
W39
W40 (Option 001 only)
W42
W69 (All mo dels except 83624B )
W73 (All mo dels except 83623B /L and 83624B )

Figure 7-35. A20 RF Interface
Agilent 8360 Disassembly and Replacement Procedures
RF Deck

F or the follo wing pro cedures, refer to Figure 7-36 through Figure 7-50 for
cable lo cations and for screw lo cations.
a. Most micro circuits ha v e a cable connecting them to the A20 RF in terface assembly .
Disconnect the cable from the A20 assembly (if y ou ha v e already remov ed the A20
assembly , this step is not necessary).
b. Disconnect all semi-rigid and exible cables from the micro circuit. Be sure to lo osen
all semi-rigid cables at their other connections also. If y ou do not, the cables may b e
damaged.
c. Remo v e the micro circuits b y removing the screws that hold them to the RF dec k.
7-44 Disassembly and Replacement Procedures
RF Deck


Figure 7-36. RF Deck Cable Locations { 83640B/L and 83650B/L
Agilent 8360 Disassembly and Replacement Procedures


Figure 7-37.
RF Deck Cable Locations {
83640B/L Option 001 and 83650B/L Option 001
7-46 Disassembly and Replacement Procedures


Figure 7-38. RF Deck Cable Locations { 83640B/50B Option 006
Agilent 8360 Disassembly and Replacement Procedures
RF Deck Cables · Option 006 · RF Deck

Options 001 and 006

Figure 7-39. RF Deck Cable Locations { 83640B/50B Options 001 and 006
7-48 Disassembly and Replacement Procedures
RF Deck Cables · RF Deck


Figure 7-40. RF Deck Cable Locations { 83620B/22B and 83630B/L
Agilent 8360 Disassembly and Replacement Procedures


Figure 7-41. RF Deck Cable Locations { 83620B/22B and 83630B/L Option 001
7-50 Disassembly and Replacement Procedures


Figure 7-42. RF Deck Cable Locations { 83620B/22B/30B Option 006
Agilent 8360 Disassembly and Replacement Procedures
RF Deck Cables · Option 006 · RF Deck

Options 001 and 006

Figure 7-43. RF Deck Cable Locations { 83620B/22B/30B Options 001 and 006
7-52 Disassembly and Replacement Procedures
RF Deck Cables · RF Deck


Figure 7-44. RF Deck Cable Locations { 83623B
Agilent 8360 Disassembly and Replacement Procedures
RF Deck Cables · No Options · RF Deck


Figure 7-45. RF Deck Cable Locations { 83623B Option 001
7-54 Disassembly and Replacement Procedures
RF Deck Cables · Option 001 · RF Deck


RF Deck Cable Locations { 83623L
Agilent 8360 Disassembly and Replacement Procedures
RF Deck Cables · No Options · RF Deck


RF Deck Cable Locations { 83623L Option 001
7-56 Disassembly and Replacement Procedures
RF Deck Cables · Option 001 · RF Deck


Figure 7-46. RF Deck Cable Locations { 83624B
Agilent 8360 Disassembly and Replacement Procedures
RF Deck Cables · No Options · RF Deck


Figure 7-47. RF Deck Cable Locations { 83624B Option 001
7-58 Disassembly and Replacement Procedures
RF Deck Cables · Option 001 · RF Deck


Figure 7-48. RF Deck Cable Locations { 83623B/24B Option 006
Agilent 8360 Disassembly and Replacement Procedures
RF Deck Cables · Option 006 · RF Deck

Options 001 and 006

Figure 7-49. RF Deck Cable Locations { 83623B/24B Options 001 and 006
7-60 Disassembly and Replacement Procedures
RF Deck Cables · RF Deck


Figure 7-50. RF Deck Screw Locations
Agilent 8360 Disassembly and Replacement Procedures
RF Deck

Motherboard Disassembly and Reassembly
T-15 T orx Screwdriv er
#2 x 4 P ozidriv Screwdriv er
5/16 in. Op en End W renc h
5mm Ball and So c k et Driver
T o disassemble the motherb oard, follo w this pro cedure in the order given.
T o reassemble the motherb oard, follo w this pro cedure in the rev erse order.
Cable routing may a ect instrumen t p erformance. Refer to Figure 7-1 for a
diagram of the instrumen t cable routing.
Refer to the o w c hart in Figure 7-51 to determine the necessary disassembly steps for a given
assembly .
7-62 Disassembly and Replacement Procedures
T ools Required · Description and Procedure · Motherboard


Figure 7-51. Motherboard Disassembly Flo w chart
Agilent 8360 Disassembly and Replacement Procedures
Motherboard

Y ou can easily blow the fuse for the 13.5 V supply on the preregulator
assembly if y ou do not disconnect the A C p ow er cord. Also, the danger of
sho c k is increased if the p ow er cord is connected.
screwdriver in either slot in the trim strip and lift to remov e, see Figure 7-52).
An RF braid is in eac h of the side c hannels on the top and b ottom of the
instrumen t. When y ou reassemble the instrumen t, b e sure the braids remain
in the side co v er c hannels.
Figure 7-52. Front Edge Trim Strip
7-64 Disassembly and Replacement Procedures
Motherboard

Refer to Figure 7-53.
Figure 7-53. Front P anel Screw Locations
Agilent 8360 Disassembly and Replacement Procedures
Motherboard

Refer to Figure 7-54.
Instrumen ts with Option 004 do not ha v e all the cables sho wn.
002), disconnect the four coaxial cables (W6, W7, W8, and W9) attached to the BNC
connectors, and the source mo dule in terface cable (W10).
Figure 7-54. Front P anel Coaxial and Ribbon Cable Locations
7-66 Disassembly and Replacement Procedures
Motherboard

Refer to Figure 7-55.
T o av oid damaging the connector, do not disassemble the RF output
connector assembly . Only unscrew the connector on the end of the cable.
assembly .

Figure 7-55. Remo ving RF Output Connector (Front P anel)
Agilent 8360 Disassembly and Replacement Procedures
Motherboard

When y ou re-attach the connector and cable to the syn thesizer, rst, lo osen
the n ut b et w een cable W43 and the connector. Then rev erse the steps ab ov e.
Finally , torque the n ut b et w een cable W43 and the connector J1 to 10 in-lb
(112 N-cm).
7-68 Disassembly and Replacement Procedures
Motherboard

Refer to Figure 7-56.
T o av oid damaging the connector, do not disassemble the RF output
connector assembly . Only unscrew the connector on the end of the cable.
Figure 7-56. Remo ving RF Output Connector (Option 004)
Agilent 8360 Disassembly and Replacement Procedures
Motherboard

Refer to Figure 7-57.
con tinuing.
Figure 7-57. Rear P anel Remo v al
7-70 Disassembly and Replacement Procedures
Motherboard

Refer to Figure 7-58.
mo dule in terface.
the source mo dule in terface cable from the A19 assembly .
Figure 7-58. Rear P anel Cable Locations
Agilent 8360 Disassembly and Replacement Procedures
Motherboard

Refer to Figure 7-59.
\7. J1 RF Output Connector Remo v al", and \5. Rear P anel Remo v al" b efore con tinuing.
All cables are lab eled for easy reassembly . F or iden ti cation purp oses, the
extractors on the PC assemblies are color-co ded to indicate their assembly
n um b er.

Figure 7-59. Major Assemblies
7-72 Disassembly and Replacement Procedures
Motherboard

Refer to Figure 7-60.
con tinuing.

Figure 7-60. W51 Location
Av oid damage to the v arious ribb on cables and their connectors attached to
the RF dec k assembly when removing that assembly in the next step. Mov e
all exible coaxial cables out of the w ay of the RF dec k assembly comp onen ts.
Agilent 8360 Disassembly and Replacement Procedures
Motherboard

Refer to Figure 7-61.
\7. J1 RF Output Connector Remo v al", \5. Rear P anel Remo v al", \6. Ma jor Assem blies
Remo v al", and \7. RF Dec k Remo v al" b efore con tinuing.

Figure 7-61. Motherboard Cables
7-74 Disassembly and Replacement Procedures
Motherboard


Figure 7-62. Motherboard Screws
Agilent 8360 Disassembly and Replacement Procedures
Motherboard



T able 8-1.
Adjustments and P erformance T ests Required
After Repair or Replacement of an Assembly
Assembly Adjustmen t Performance T est
A1 F ront P anel Keyb oard None F ull Self{T est
F ront P anel Self{T est
A2 Source Mo dule Interface None F ull Self{T est
A3 F ront P anel Pro cessor None F ull Self{T est
F ront P anel Self{T est
A4 F ractional{N None F ull Self{T est
F requency Switching Time
A5 YO Phase Detector YO Lo op Gain F ull Self{T est
F requency Switching Time
Single Sideband Phase Noise
A6 Sampler Sampler Assem bly F ull Self{T est
F requency Switching Time
Single Sideband Phase Noise
This adjustment or p erformance test is required only for the Agilent 8360 B-Series sw ept signal
generator. Do not p erform this adjustment or test on the 8360 L-Series sw ept CW generator.
The adjustmen ts listed are required only after r ep air of the assembly . The r eplac ement assembly
is adjusted at the factory and is not instrument-dep enden t.
Agilent 8360 P ost{Repair 8-1

T able 8-1.
Adjustments and P erformance T ests Required
After Repair or Replacement of an Assembly (continued)
Assembly Adjustment Performance T est
A7 Reference None F ull Self{T est
F requency Switching Time
Single Sideband Phase Noise
A8 Mo dulation Generator AM/FM D A C O set and Gain
F ull Self{T est
Mo dulation Generator Flatness
A9 Pulse A9 Pulse Bd Jump er
F ull Self{T est
Maximum Leveled P ow er
Low P ow er SRD Bias
A10 ALC Mo dulator O set and Gain
F ull Self{T est
ALC P ow er Level Accuracy
P ow er Accuracy
P ow er Flatness
Maximum Leveled P ow er
A11 FM Driver FM Gain
F ull Self{T est
A12 Multiplier/Filter Driver Amp/Mult Adjustmen ts
F ull Self{T est
Maximum Leveled P ow er
This adjustment or p erformance test is required only for the 8360 B-Series sw ept signal
generator. Do not p erform this adjustment or test on the 8360 L-Series sw ept CW generator.
The adjustments listed are required only after r ep air of the assembly . The r eplac ement assembly
is adjusted at the factory and is not instrument-dep enden t.
8-2 P ost{Repair Agilent 8360

T able 8-1.
Adjustments and P erformance T ests Required
After Repair or Replacement of an Assembly (continued)
Assembly Adjustmen t Performance T est
A13 YO Driver YO Driver +10 V Reference
F ull Self{T est
YO Driver Gain and Linearity
Sw ept F requency Accuracy
F requency Switching Time
Single Sideband Phase Noise
A14 Sw eep Generator Sw eep Ramp F ull Self{T est
Sw ept F requency Accuracy
A15 CPU Selftest P atches (L Mo dels only)
F ull Self{T est
A16 P ostregulator None F ull Self{T est
A17 Recti er/Filter None F ull Self{T est
A18 Switching Regulator None F ull Self{T est
A19 Rear P anel Interface None F ull Self{T est
A20 RF Interface None F ull Self{T est
Op eration V eri cation
A21 Alphanumeric Displa y None F ull Self{T est
A22 Motherb oard None F ull Self{T est
Op eration V eri cation
A23 10 MHz Reference
10 MHz Standard F ull Self{T est
Single Sideband Phase Noise
This adjustment or p erformance test is required only for the 8360 B-Series sw ept signal
generator. Do not p erform this adjustment or test on the 8360 L-Series sw ept CW generator.
The adjustmen ts listed are required only after r ep air of the assembly . The r eplac ement assembly
is adjusted at the factory and is not instrument-dep enden t.
Agilent 8360 P ost{Repair 8-3

T able 8-1.
Adjustments and P erformance T ests Required
After Repair or Replacement of an Assembly (continued)
Assembly Adjustment Performance T est
A24 Low Band Mo dulator O set and Gain
F ull Self{T est
ALC P ow er Level Accuracy
P ow er Accuracy
P ow er Flatness
P ow er Flatness
Maximum Leveled P ow er
Single Sideband Phase Noise
Pulse Mo dulation On/O
Pulse Mo dulation Video
A26 YIG Oscillator YO Driver Gain and Linearity
F ull Self{T est
Sw ept F requency Accuracy
F requency Switching Time
Maximum Leveled P ow er
Single Sideband Phase Noise
A28 Amp/Mult Amp/Mult Adjustmen ts
F ull Self{T est
Maximum Leveled P ow er
A29 Amp/Filter Amp/Filter Adjustmen ts F ull Self{T est
Maximum Leveled P ow er
A30 Directional Coupler ALC P ow er Level Accuracy
F ull Self{T est
P ow er Flatness
P ow er Accuracy
Maximum Leveled P ow er
This adjustment or p erformance test is required only for the 8360 B-Series sw ept signal
generator. Do not p erform this adjustment or test on the 8360 L-Series sw ept CW generator.
The adjustments listed are required only after r ep air of the assembly . The r eplac ement assembly
is adjusted at the factory and is not instrument-dep enden t.
8-4 P ost{Repair Agilent 8360

T able 8-1.
Adjustments and P erformance T ests Required
After Repair or Replacement of an Assembly (continued)
Assembly Adjustmen t Performance T est
A30 Bridge Detector ALC P ow er Level Accuracy
F ull Self{T est
P ow er Flatness
P ow er Accuracy
Maximum Leveled P ow er
A31 Step A tten uator ALC P ow er Level Accuracy
F ull Self{T est
Step A tten uator Flatness
Maximum Leveled P ow er
A tten uator V eri cation
(only if data is required)
A32 Doubler P ow er Flatness F ull Self{T est
P ow er Accuracy
P ow er Flatness
Maximum Leveled P ow er
A33 Ampli er/Switch P ow er Flatness F ull Self{T est
Maximum Leveled P ow er
Pulse Mo dulation On/O
A33 Ampli er P ow er Flatness F ull Self{T est
Maximum Leveled P ow er
Pulse Mo dulation On/O
A34 Rear P anel BNC None F ull Self{T est
A38 Dual Mo dulator Amp/Filter Adjustments
Maximum leveled P ow er
A38 Mo d Doubler Amp Mo dulator O set and Gain
ALC P ow er Level Accuracy
Pulse Mo dulation On/O
P ow er Flatness
Pulse Mo dulation Video
This adjustment or p erformance test is required only for the 8360 B-Series sw ept signal
generator. Do not p erform this adjustment or test on the 8360 L-Series sw ept CW generator.
The adjustmen ts listed are required only after r ep air of the assembly . The r eplac ement assembly
is adjusted at the factory and is not instrument-dep enden t.
The entire ampli er/ lter adjustment pro cedure do es not need to b e p erformed. P erform only
the squegging clamp p ortions of this adjustment.
Agilent 8360 P ost{Repair 8-5

T able 8-1.
Adjustments and P erformance T ests Required
After Repair or Replacement of an Assembly (continued)
Assembly Adjustment Performance T est
A39 Directional Coupler None F ull Self{T est
Maximum Leveled P ow er
A41 Dio de Switch None F ull Self{T est
Maximum Leveled P ow er
A42 Pulse Mo d Switched Filter Mo dulator O set and Gain
Maximum Leveled P ow er
Pulse Mo dulation On/O
Pulse Mo dulation Video
B1 F an None Spurious Signals
CR1 Detector ALC P ow er Level Accuracy
F ull Self{T est
P ow er Flatness
P ow er Accuracy
Maximum Leveled P ow er
FL1 Line Mo dule None Spurious Signals
FL2 8 GHz LPF Sampler Assem bly F ull Self{T est
J1 RF Output Connector P ow er Flatness (automated)
F ull Self{T est
Step A tten uator (automated)
Maximum Leveled P ow er
RPG1 Rotary Pulse Generator None F ull Self{T est
F ront P anel Self{T est
T1 T oroid P ow er T ransformer None F ull Self{T est
This adjustment or p erformance test is required only for the 8360 B-Series sw ept signal
generator. Do not p erform this adjustment or test on the 8360 L-Series sw ept CW generator.
The adjustments listed are required only after r ep air of the assembly . The r eplac ement assembly
is adjusted at the factory and is not instrument-dep enden t.
8-6 P ost{Repair Agilent 8360

Replaceable parts include ma jor assemblies and all c hassis hardw are. T able 9-1 lists reference
designations and abbreviations used in this c hapter.
F or information on removing and replacing assemblies, see Chapter 7, \Disassembly and
Replacemen t Pro cedures."
Many ma jor assemblies are co v ered b y the mo dule{exc hange program. Under the terms of
the program, factory{repaired and tested assemblies are av ailable on a trade{in basis (a
defectiv e assembly must b e returned for credit). Exc hange assemblies meet all new assembly
sp eci cations, but are less exp ensiv e. Figure 9-1 illustrates the mo dule exc hange pro cedure.
In order to prev en t damage when returning pro ducts to Agilent T ec hnologies, use original or
comparable pac k aging.
Order spare assembly sto c k using the new assembly part n um b er.
If y ou ha v e an y questions, con tact y our Agilent T ec hnologies customer engineer for the latest
information ab out this program.
The follo wing information is given in the tables in this c hapter:
Agilent part n um b er.
P art quan tity for that gure. There may b e more of that part in other gures.
P art description (see T able 9-1 for abbreviations).
Agilent 8360 Replaceable P arts 9-1

F or an y listed part, request the Agilent part n um b er and quan tity required from the nearest
Agilent oce.
Ho w to Order P arts . . . F ast!
Agilent parts sp ecialists ha v e direct on{line access to the replaceable parts listed in this
man ual. F our{day delivery is standard; there is a c harge for hotline (one{day) delivery .
In the United States, call the follo wing toll{free n um b er: (800) 227{8164
Monday through F riday , 6 am to 5 pm (Paci c Standard Time)
Outside the United States, con tact y our nearest Agilent oce.
9-2 Replaceable P arts Agilent 8360
Ordering Information

T able 9-1. Reference Designations and Abbreviations
A Amp erage; Assem bly
A T T ermination
LPF Low P ass Filter
A Y Assem bly
M Male; Maximum; Meter; Mil; Milli
B F an; Motor
P AN{HD P an Head
CMPNT Comp onen t
RP Rear P anel
RPG Rotary Pulse Generator
SKT So ck et
SRD Step Recovery Dio de
T T ransformer Thickness; T ap er;
FP F ront P anel
HD Hand; Hard; Heavy Duty; Head
TR{HD T russ Head
HEX Hexadecimal; Hexagon; Hexagonal
UL Underwriters Lab oratories
YIG Yttrium Iron Garnate
J Electrical Connector (Stationary
Agilent 8360 Replaceable P arts 9-3


Figure 9-1. Module Exchange Program
9-4 Replaceable P arts Agilent 8360

Major Assemblies{T op View

Figure 9-2. Major Assemblies (1 of 5)
Agilent 8360 Replaceable P arts 9-5


Figure 9-2. Major Assemblies (2 of 5)
9-6 Replaceable P arts Agilent 8360
Major Assemblies{RF Deck


Figure 9-2. Major Assemblies (3 of 5)
Agilent 8360 Replaceable P arts 9-7
Major Assemblies{RF Deck


Figure 9-2. Major Assemblies (4 of 5)
9-8 Replaceable P arts Agilent 8360
Major Assemblies{RF Deck


Figure 9-2. Major Assemblies (5 of 5)
Agilent 8360 Replaceable P arts 9-9
Major Assemblies{RF Deck

T able 9-2. Major Assemblies
Ref.
Desig.
A1 08360-60001 1 FRONT P ANEL KEYBO ARD ASSEMBL Y
A2 08360-60002 1 SOUR CE MODULE INTERF A CE ASSEMBL Y
A3 06360-60003 1 FRONT P ANEL PR OCESSOR ASSEMBL Y
A5 08360-60209 1 YO LOOP ASSEMBL Y
A5 08360-69209 YO LOOP ASSEMBL Y (Rebuilt{Exchange)
A6 08360-69214 SAMPLER ASSEMBL Y (Rebuilt{Exchange)
A7 08360-69008 REFERENCE ASSEMBL Y (Rebuilt{Exchange)
A8 08360-60166 1 MODULA TION GENERA TION ASSEMBL Y (Option 002)
(Agilent 8360 B-series only)
A8 08360-69166 MODULA TION GENERA TION ASSEMBL Y (Rebuilt{Exchange)
A8W1 1258-0124 1 JUMPER{AM INPUT IMPED ANCE (8360 B-series only)
A8W2 1258-0124 1 JUMPER{FM INPUT IMPED ANCE (8360 B-series only)
A9 08360-60294 1 PULSE ASSEMBL Y REPLA CEMENT KIT (Contains 08360-60292 and
A10 08360-60222 1 ALC ASSEMBL Y (Agilent 83620B/22B/24B)
A10 08360-69222 ALC ASSEMBL Y (Rebuilt{Exchange)
A10 08360-60206 1 ALC ASSEMBL Y (Agilent 83640B/L and Agilent 83650B/L/23L/30L)
A10 08360-69206 ALC ASSEMBL Y (Rebuilt{Exchange)
A11 08360-60011 1 FM DRIVER ASSEMBL Y (8360 B-series only)
A11W1 8159-0005 1 ZER O OHM RESISTOR{FM INPUT IMPED ANCE (8360 B-series only)
9-10 Replaceable P arts Agilent 8360

T able 9-2. Major Assemblies (continued)
Ref.
Desig.
A12 08360-60212 1 MUL TIPLIER/FIL TER DRIVER ASSEMBL Y
A12 08360-69212 MUL TIPLIER/FIL TER DRIVER ASSEMBL Y (Rebuilt{Exchange)
A13 08360-69207 YO DRIVER ASSEMBL Y (Rebuilt{Exc hange)
A14 08360-69014 SWEEP GENERA TOR ASSEMBL Y (Rebuilt{Exchange)
A15 08360-69264 CPU ASSEMBL Y (Rebuilt{Exchange)
A15BT1 1420-0394 1 BA TTER Y (Used with CPU Assem bly 08360-60264)
A16 08360-69016 POSTREGULA TOR ASSEMBL Y (Rebuilt-Exchange)
0400-0018 1 GR OMMET-CHAN .052 TH (P art of A16)
A16F1 2110-0047 1 FUSE (INCH) 1A 125V NTD BI
A16F2 2110-0618 3 FUSE (INCH) 5A 125V NTD BI
A16F3 2110-0713 2 FUSE{SUBMINIA TURE 10A 125V NTD AX
A16F5 2110-0618 FUSE (INCH) 5A 125V NTD BI
A16F6 2110-0618 FUSE (INCH) 5A 125V NTD BI
A16F7 2110-0425 2 FUSE (INCH) 2A 125V NTD BI
A17 08360-60017 1 RECTIFIER/FIL TER ASSEMBL Y
A17 08360-69017 RECTIFIER/FIL TER ASSEMBL Y (Rebuilt{Exchange)
A17F1 2110-0303 1 FUSE (INCH) 2A 250V TD FE UL
A18 08360-69198 SWITCHING REGULA TOR ASSEMBL Y (Rebuilt{Exchange)
A18F1 2110-0201 1 FUSE (INCH) 0.25A 250V TD FE UL
Agilent 8360 Replaceable P arts 9-11
P art · Qty Description · Num b er

T able 9-2. Major Assemblies (continued)
Ref.
Desig.
A19 08360-60019 1 REAR P ANEL INTERF A CE ASSEMBL Y
A20 08360-60266 1 RF INTERF A CE ASSEMBL Y
A20F1 2110-0425 1 FUSE (INCH) 2A 125V NTD BI
A22 08360-60196 1 MOTHERBOARD ASSEMBL Y
A23 10811E 1 10 MHZ REFERENCE ST AND ARD (Includes W30)
A23 1813-1258 1 OCX O 10 MHz
A24 5086-6463 LO WBAND ASSEMBL Y (Rebuilt{Exc hange)
A26 5086-7515 1 YIG OSCILLA TOR (2.0{7.5 GHz)
A26 5086-6515 YIG OSCILLA TOR (2.0{7.5 GHz) (Rebuilt{Exc hange)
A28 5086-7622 1 AMPLIFIER/MUL TIPLIER
A28 5086-6622 AMPLIFIER/MUL TIPLIER (Rebuilt{Exchange)
A29 5086-7616 1 FIL TER (Agilent 83620B/22B/23B/24B)
A29 5086-6616 FIL TER (Rebuilt{Exchange)
A29 5086-7623 1 AMPLIFIER/FIL TER
(Agilent 83630B/L, 83640B/L, and 83650B /L)
A29 5086-6623 AMPLIFIER/FIL TER (Rebuilt{Exchange)
A29 5086-7636 1 AMPLIFIER/FIL TER (83630B /40B/50B Option 006)
A29 5086-6636 AMPLIFIER/FIL TER (Rebuilt{Exchange)
A30 5086-7645 1 BRIDGE DETECTOR (83620B/22B and 83630B /L)
A30 0955-0148 1 W A VE DIRECTIONAL COUPLER 20 GHZ MAX
A30 5086-7644 1 BRIDGE DETECTOR (83640B/L and 83650B/L)
A30 5086-7656 1 BRIDGE DETECTOR (83620B/22B/30B Option 006)
9-12 Replaceable P arts Agilent 8360

T able 9-2. Major Assemblies (continued)
Ref.
Desig.
(83620B/22B/24B, 83623B/L, and 83630B/L)
A31 33326-60005 1 STEP A TTENUA TOR (83650B/L)
A32 5086-7607 1 DOUBLER (83640B/L)
A32 5086-6607 DOUBLER (Rebuilt{Exchange)
A32 5086-6604 DOUBLER (Rebuilt{Exchange)
A33 5086-6465 1 AMPLIFIER/SWITCH (Rebuilt{Exchange)
A33 5086-7510 1 AMPLIFIER (2{20 GHz) (83624B)
A33 5086-7610 1 AMPLIFIER (2{20 GHz) (Rebuilt{Exchange)
A34 08360-60061 1 RP FLOA TING BNC ASSEMBL Y (Option 004)
A34W1 08360-60099 1 EXTERNAL LEVELING A34J3/A10J4 (Option 004)
A34W2 08360-60100 1 AM INPUT A34J1/A10J8 (Option 004)
A34W3 08360-60101 1 FM INPUT A34J2/A11J1 (Option 004)
A34 08360-60186 1 RP FLOA TING BNC ASSEMBL Y (Option 002 and 004)
A34W1 08360-60099 1 EXTERNAL LEVELING A34J3/A10J4 (Option 002 and 004)
A34W2 08360-60178 1 AM INPUT A34J1/A8J7 (Option 002 and 004)
A34W3 08360-60179 1 FM INPUT A34J2/A8J10 (Option 002 and 004)
A36 NOT USED
A38 5086-7602 1 DUAL MODULA TOR (B Mo del)
A38 5086-7634 1 AM MODULA TOR (8360 B-series Option 006 and 83623L)
A38 5087-7017 1 Mo d Doubler Amp (83630L/40L/50L)
A40 11742A 1 DC BLOCKING CAP A CITOR (83623L )
A41 5086-7327 1 DIODE SWITCH (All mo dels except 83624B)
Agilent 8360 Replaceable P arts 9-13
P art · Qty Description · Num b er

T able 9-2. Major Assemblies (continued)
Ref.
Desig.
A42 5086-7628 1 PULSE MODULA TION SWITCHED FIL TER (83620B/22B
A42 5086-7629 1 PULSE MODULA TION SWITCHED FIL TER (83623B/24B
A42 5086-7630 1 PULSE MODULA TION SWITCHED FIL TER (83630B Option 006)
A42 5086-7631 1 PULSE MODULA TION SWITCHED FIL TER (83640B/50B
A T2 1810-0118 1 TERMINA TION SMA MALE 50
A T3 0955-0216 1 A TTENUA TOR 20 DB
CR1 08360-60147 1 DETECTOR REPLA CEMENT KIT (83623B/L and 83624B)
F1 2110-0227 1 FUSE (INCH) 5A 250V NTD FE UL (See T able 9-7 )
F1 2110-0003 1 FUSE (INCH) 3A 230V NTD FE UL (See T able 9-7 )
FL2 0955-0485 1 8 GHZ LO W P ASS FIL TER
(83620B/22B/24B, 83623B /L, and 83630B/L)
J1 5063-1700 1 RF OUTPUT CONNECTOR (83640B/L and 83650B/L) { F or serial
pre xes 4040A, also order W43
1250-1158 1 AD APTER SMA(F)|SMA(F) (83623L) USED WITH A40 DC
9-14 Replaceable P arts Agilent 8360


Figure 9-3. Cables (1 of 19)
Agilent 8360 Replaceable P arts 9-15
Cables{Front P anel


Figure 9-3. Cables (2 of 19)
9-16 Replaceable P arts Agilent 8360
Cables{T op View

8360 B-Series Option 002

Figure 9-3. Cables (3 of 19)
Agilent 8360 Replaceable P arts 9-17
Cables{T op View

Cables{83640B/L and 83650B/L RF Deck

Figure 9-3. Cables (4 of 19)
9-18 Replaceable P arts Agilent 8360
No Options

Cables{83640B/L and 83650B/L RF Deck

Figure 9-3. Cables (5 of 19)
Agilent 8360 Replaceable P arts 9-19
Option 001


Figure 9-3. Cables (6 of 19)
9-20 Replaceable P arts Agilent 8360
Cables{83640B/50B RF Deck · Option 006

Options 001 and 006

Figure 9-3. Cables (7 of 19)
Agilent 8360 Replaceable P arts 9-21
Cables{83640B/50B RF Deck

Cables{83620B/22B and 83630B/L RF Deck

Figure 9-3. Cables (8 of 19)
9-22 Replaceable P arts Agilent 8360
No Options

Cables{83620B/22B and 83630B/L RF Deck

Figure 9-3. Cables (9 of 19)
Agilent 8360 Replaceable P arts 9-23
Option 001


Figure 9-3. Cables (10 of 19)
9-24 Replaceable P arts Agilent 8360
Cables{83620B/22B/30B RF Deck · Option 006

Options 001 and 006

Figure 9-3. Cables (11 of 19)
Agilent 8360 Replaceable P arts 9-25
Cables{83620B/22B/30B RF Deck


Figure 9-3. Cables (12 of 19)
9-26 Replaceable P arts Agilent 8360
Cables{83623B RF Deck · No Options


Figure 9-3. Cables (13 of 19)
Agilent 8360 Replaceable P arts 9-27
Cables{83623B RF Deck · Option 001


Figure 9-3. Cables (14 of 19)
9-28 Replaceable P arts Agilent 8360
Cables{83624B RF Deck · No Options


Figure 9-3. Cables (15 of 19)
Agilent 8360 Replaceable P arts 9-29
Cables{83624B RF Deck · Option 001


Figure 9-3. Cables (16 of 19)
9-30 Replaceable P arts Agilent 8360
Cables{83623B/24B RF Deck · Option 006

Options 001 and 006

Figure 9-3. Cables (17 of 19)
Agilent 8360 Replaceable P arts 9-31
Cables{83623B/24B RF Deck

Figure 9-3. Cables (18 of 19)
9-32 Replaceable P arts Agilent 8360
Cables{83623L RF Deck · No Options


Figure 9-3. Cables (19 of 19)
Agilent 8360 Replaceable P arts 9-33
Cables{83623L RF Deck · Option 001

T able 9-3. Cables
Ref.
Desig.
W2 08360-60056 1 SMI/MOTHERBO ARD{A2J1/A22J1
W3 08360-60057 1 F.P . PR OCESSOR/MOTHERBO ARD A3J2/A22J2
W4 08360-60055 1 F.P . PR OCESSOR/DISPLA Y A3J3/A21J2
W6 08360-60070 1 EXTERNAL PULSE J2/A9J4 (8360 B-Series)
W6 08360-60094 1 EXTERNAL PULSE J8/A9J4 (8360 B-Series Option 004)
W6 08360-60172 1 EXTERNAL PULSE J2/A8J3 (8360 B-Series Option 002)
W6 08360-60175 1 EXTERNAL PULSE J8/A8J3 (8360 B-Series Option 002 and 004)
W7 08360-60068 1 AM INPUT J3/A10J8 (8360 B-Series)
W7 08360-60173 1 AM INPUT J3/A8J7 (8360 B-Series Option 002)
W8 08360-60069 1 FM INPUT J4/A11J1 (8360 B-Series)
W8 08360-60174 1 FM INPUT J4/A8J10 (8360 B-Series Option 002)
W10 08360-60092 1 FP SMI LEVELING{A2J2/A10J2
W11 08360-60301 1 10 MHZ REF{A7J5/A4J1
W12 08360-60302 1 30{60 MHZ{A4J3/A5J2
W13 08360-60076 1 SAMPLER IF{A6J2/A5J1
W14 08360-60083 1 HIGH FREQ PLL{A5J3/A26J2
W15 08360-60072 1 40 MHZ REF{A7J6/A6J1
W15 08360-60182 1 40 MHz REF{A7J6/A8J1 (8360 B-Series Option 002)
W17 08360-60071 1 O VEN STD{A23J1/A7J2
W18 08360-60073 1 1 MHZ{A7J4/A14J1
W19 08360-60303 1 10 MHZ REAR P ANEL {A4J2/J7
9-34 Replaceable P arts Agilent 8360
P art · Qty Description · Num b er

T able 9-3. Cables (continued)
Ref.
Desig.
W22 08360-60228 1 PRE-PULSE LEVELING DET{A38J13/A10J6
W22 08360-60245 1 PRE-PULSE LEVELING DET{A42J4/A10J6
(8360 B-Series Option 006)
W24 08360-60107 1 LO WBAND DETECTOR{A24J3/A10J5
W25 08360-60106 1 HI BAND DETECTOR{CR1/A10J7
W26 08360-60079 1 YO FM SIGNAL A11J2/A26J3 (8360 B-Series only)
W28 08360-60098 1 TRIGGER OUTPUT{A14J2/J9
W29 08360-60031 1 PREREG IN/PREREG OUT{A18J1/A17J1
W30 08360-60063 1 MOTHERBO ARD/10MHZ STD{A22J7/A23J2
W31 08360-60037 1 REAR P ANEL/MOTHERBO ARD{A19J1/A22J8
W32 08360-60226 1 PULSE MOD HI{A22J3/A38J3
W32 08360-60247 1 SLO W PULSE{A22J3/A42J3 (8360 B- Series Option 006)
W33 08360-60088 1 PULSE MOD LO{A22J4/A24J6 (not used on 83623L/30L/40L/50L)
W35 08360-60225 1 ALC MOD HI{A22J6/A38J2
W36 08360-60230 1 RF INTERF A CE/AMP SWITCH{A20J14/A33J1 (83623B)
W36 08360-60191 1 RF INTERF A CE/DOUBLER{A20J5/A32J1 (83640B/L and 83650B/L)
W37 08360-60189 1 RF INTERF A CE/AMP MUL T{A20J9/A28J1
W38 08360-60227 1 RF INTERF A CE/AM MOD{A20J1/A38J1
(8360 B-Series Option 006)
W40 08360-60045 1 RF INTERF A CE/A TTEN A20J6/A31J1 (Option 001)
Agilent 8360 Replaceable P arts 9-35
P art · Qty Description · Num b er

T able 9-3. Cables (continued)
Ref.
Desig.
W41 08360-60190 1 RF INTERF A CE/AMP FIL TER{A20J8/A29J1
W43 08360-20072 1 DIR COUPLER/RF OUT{A30J2/J1 (83620B/22B/24B, 83623B/L, and
W43 08360-20072 1 A TTEN/RF OUT{A31J2/J1 (83620B/22B/24B, 83623B/L, and
W43 08360-20073 1 DIR COUPLER/RF OUT{A30J2/J1 (83620B/22B/24B, 83623B/L, and
W43 08360-20073 1 A TTEN/RF OUT{A31J2/J1 (83620B/22B/24B, 83623B/L, and
W43 08360-20097 1 DIR COUPLER/RF OUT{A30J2/J1 (83640B/L and 83650B/L)
(83640B/L and 83650B/L Option 001)
W43 08360-20098 1 DIR COUPLER/RF OUT{A30J2/J1 (83640B/L and 83650B/L Option
W43 08360-20098 1 A TTEN/RF OUT{A31J2/J1 (83640B/L and 83650B/L Option 014)
W44 08360-20191 1 BRIDGE DET/A TTEN A30J5/A31J1 (83620B /22B/30B Options 001
W44 08360-20071 1 DIR COUPLER/A TTEN A30J2/A31J1 (83623B /24B Options 001 and
W44 08360-20193 1 BRIDGE DET/A TTEN A30J5/A31J1 (83640B /L and 83650B /L Option
W45 08360-20185 1 AMP FIL TER/BRIDGE DET{A29J3/A30J4 (83620B/22B and
W45 08360-20190 1 AMP FIL TER/BRIDGE DET{A29J3/A30J4 (83620B/22B and
W45 08360-20189 1 DC BLOCKING CAP A CITOR/DIR COUPLER{A40J2/A30J1
W45 08360-20186 1 DC BLOCKING CAP A CITOR/DIR COUPLER{A40J2/A30J1
9-36 Replaceable P arts Agilent 8360

T able 9-3. Cables (continued)
Ref.
Desig.
W47 08360-20075 1 LO WBAND/AMP SWITCH{A24J4/A33J3 (83623B )
W47 08360-20163 1 LO WBAND/AMP FIL TER{A24J4/A29J2 (83620B/22B, 83623L, and
W47 08360-20139 1 LO WBAND/DOUBLER{A24J4/A32J3 (83640B/L and 83650B/L)
W47 08360-20246 1 LO WBAND/PLS MOD SW FIL T{A24J4/A42J7 (83620B/23B/30B
W48 08360-20236 1 AMP MUL T/A UX OUTPUT{A28J4/J14 (Option 004)
W49 08360-20171 1 DIODE SWITCH/AMP MUL T{A41J3/A28J5 (All mo dels except
W49 08360-20230 1 COUPLER/AMP MUL T{A39J2/A28J5 (83624B)
W50 08360-20169 1 DIODE SWITCH/LO WBAND{A41J2/A24J5 (All mo dels except
W51 08360-20173 1 A TTENUA TOR/LPF{A T3J2/FL2J1
W51 08360-20172 1 A TTENUA TOR/LPF{A T3J2/FL2J1 (Option 004)
W53 NOT USED
W54 08360-20087 1 LPF/SAMPLER{FL2J2/A6J3
W55 1250-1159 1 DIR COUPLER/DIODE SWITCH{A39J2/A41J1 (All mo dels except
Agilent 8360 Replaceable P arts 9-37

T able 9-3. Cables (continued)
Ref.
Desig.
W56 08360-20258 1 AMP MUL T/AM MOD{A28J3/A38J8 (8360 B-Series Option 006),
W57 08360-20203 1 AM MOD/AMP FIL TER{A38J9/A29J2 (8360 B-Series Option 006)
W58 08360-20221 1 FIL TER/AMPLIFIER{A29J3/A33J2 (83624B)
W58 08360-20137 1 AMP FIL TER/DOUBLER{A29J3/A32J2 (83640B /L and 83650B/L)
W58 08360-20137 1 AMP FIL TER/DC BLOCKING CAP A CITOR{A29J3/A40J1
W58 08360-20250 1 AMP FIL TER/PLS MOD SW FIL T{A29J3/A42J5 (8360 B-Series
W59 08360-20070 1 AMP SWITCH/DIR COUPLER{A33J4/A30J1
W59 08360-20107 1 AMPLIFIER/DIR COUPLER{A33J3/A30J1 (83624B Option 001)
W59 08360-20194 1 DOUBLER/BRIDGE DET{A32J4/A30J4 (83640B/L and 83650B/L)
W59 08360-20192 1 DOUBLER/BRIDGE DET{A32J4/A30J4 (83640B/L and 83650B/L
W61 NOT USED
W62 08360-60176 1 PULSE VIDEO OUT{A8J4/J13 (8360 B-Series Option 002)
W63 08360-60177 1 PULSE SYNC OUT{A8J6/J12 (8360 B-Series Option 002)
W64 08360-60180 1 AM/FM OUTPUT{A8J11/J11 (8360 B-Series Option 002)
W65 08360-60181 1 40 MHZ OUT{A8J21/A6J1 (8360 B-Series Option 002)
9-38 Replaceable P arts Agilent 8360

T able 9-3. Cables (continued)
Ref.
Desig.
W66 08360-60183 1 PULSE OUT{A8J5/A9J4 (8360 B-Series Option 002)
W67 08360-60184 1 AM OUT{A8J8/A10J1 (8360 B-Series Option 002)
W68 08360-60185 1 FM OUT{A8J9/A11J1 (8360 B-Series Option 002)
W69 08360-60224 1 RF INTERF A CE/DIODE SWITCH{A20J12/A41J1 (All mo dels except
W70 08360-60248 1 F AST PULSE HI{A9J3/A42J1 (8360 B-Series Option 006)
W71 08360-20252 1 PLS MOD SW FIL T/BRIDGE DET{A42J6/A30J4
W71 08360-20249 1 PLS MOD SW FIL T/COUPLER{A42J6/A30J1 (83623B/24B Options
W71 08360-20251 1 PLS MOD SW FIL T/BRIDGE DET{A42J6/A30J4
(83620B/22B/30B Options 001 and 006)
W71 08360-20248 1 PLS MOD SW FIL T/COUPLER{A42J6/A30J1 (83623B/24B Option
W72 08360-60191 1 RF INTERF A CE/PLS MOD SW FIL T{A20J11/A42J8 (8360 B-Series
Agilent 8360 Replaceable P arts 9-39

Front P anel Hardw are

Figure 9-4. Front P anel Hardw are
9-40 Replaceable P arts Agilent 8360

T able 9-4. Front P anel Hardw are
Ref.
Desig.
1 0515-2043 7 SCREW-MA CH. M4 x 0.7 8MM-LG 90-DEG-FLH-HD
2 0515-0943 2 SCREW-MA CH. M4 x 0.7 12MM-LG
0535-0082 2 NUT-SPECIAL TY M3.5 x 1.27 THD; 8.44MM
4 0515-2043 SCREW-MA CH. M4 x 0.7 8MM-LG 90-DEG-FLH-HD
5 0515-0943 SCREW-MA CH. M4 x 0.7 12MM-LG
0535-0082 2 NUT-SPECIAL TY M3.5 x 1.27 THD; 8.44MM
4 (Includes the following)
0590-1251 NUT-SPCL Y 15/32-32-THD .1-IN-THK .562-WD
7 J3 FL TG BNC A Y-FM (8360 B-Series)
0590-1251 NUT-SPCL Y 15/32-32-THD .1-IN-THK .562-WD
8 J4 FL TG BNC A Y-AM (8360 B-Series)
0590-1251 NUT-SPCL Y 15/32-32-THD .1-IN-THK .562-WD
9 J5 FL TG BNC A Y-EXT (8360 B-Series)
0590-1251 NUT-SPCL Y 15/32-32-THD .1-IN-THK .562-WD
10 08360-40007 1 DISPLA Y FIL TER
11 1251-5436 4 SCREW LOCK KIT-AMP HD CONN
J1 RF OUTPUT CONNECTOR - See T able 9-2
RPG1 08340-60326 1 R OT AR Y PULSE GENERA TOR (RPG)
2190-0016 1 W ASHER-LK INTL T 3/8 IN .377-IN-ID
Agilent 8360 Replaceable P arts 9-41

Front P anel, Casting and Keypad

Figure 9-5. Front P anel, Casting and Keypad
T able 9-5. Front P anel, Casting and Keypad
Ref.
Desig.
2 08360-20051 1 FRONT P ANEL CASTING - FEEDS (8360 B-Series only)
3 06360-20050 1 FRONT P ANEL CASTING - NO FEEDS (Option 004 for all 8360 B/L
4 06360-20287 1 FRONT P ANEL CASTING - (8360 L-Series only)
F or 2, 3, and 4, also order front panel nameplate from T able 9-16.
9-42 Replaceable P arts Agilent 8360

Front P anel, A ttaching Hardw are

Figure 9-6. Front P anel, A ttaching Hardw are (1 of 3)
T able 9-6. Front P anel, A ttaching Hardw are
Ref.
Desig.
1 0515-0372 27 SCREW-MA CHINE ASSEMBL Y M3 x 0.5 8MM-LG
2 0515-0664 2 SCREW-MA CHINE ASSEMBL Y M3 x 0.5 12MM-LG
3 08360-00034 2 MOUNTING BRA CKET - DISPLA Y
4 0515-0372 SCREW-MA CHINE ASSEMBL Y M3 X 0.5 8MM-LG
5 0515-0372 SCREW-MA CHINE ASSEMBL Y M3 X 0.5 8MM-LG
Agilent 8360 Replaceable P arts 9-43
P art · Qty Description · Num b er

Front P anel, A ttaching Hardw are

Figure 9-6. Front P anel, A ttaching Hardw are (2 of 3)
9-44 Replaceable P arts Agilent 8360

Front P anel, A ttaching Hardw are

Figure 9-6. Front P anel, A ttaching Hardw are (3 of 3)
Agilent 8360 Replaceable P arts 9-45

Rear P anel Hardw are

Figure 9-7. Rear P anel Hardw are
9-46 Replaceable P arts Agilent 8360

Rear P anel Hardw are
T able 9-7. Rear P anel (1 of 2)
Ref.
Desig.
2 0515-1372 5 SCREW-MA CHINE ASSEMBL Y M3 x 0.5 8MM-LG
3 BNC CONNECTORS (Order A34 of T able 9-2 )
4 08360-60094 1 J8 - CA A Y-PLS INPUT (Option 004)
6960-0041 1 PLUG-HOLE TR-HD FOR .5-D-HOLE NYL (Standard)
5 J6 - CA A Y-10MHZ IN-See T able 9-3
2190-0102 4 W ASHER LK INTL T 15/32 IN .472-IN-ID
6 J10 - CA A Y-V OL TS/GHZ-See T able 9-3
2190-0102 W ASHER LK INTL T 15/32 IN .472-IN-ID
7 J7 - CA A Y-10MHZ OUT-See T able 9-3
2190-0102 W ASHER LK INTL T 15/32 IN .472-IN-ID
8 J9 - CA A Y-ANAL INTFC-See T able 9-3
2190-0102 W ASHER LK INTL T 15/32 IN .472-IN-ID
9 0515-0380 8 SCREW-MA CHINE ASSEMBL Y M4 x 0.7 10MM-LG
11 1251-5436 4 SCREW LOCK KIT-AMP HD CONN
12 0515-1037 2 SCREW-MA CHINE ASSEMBL Y M3 x 0.5 45MM-LG
13 0515-0372 9 SCREW-MA CHINE ASSEMBL Y M3 x 0.5 8MM-LG
14 LINE MODULE - FIL TERED (See T able 9-2 )
15 08360-60032 1 A C V OL T A GE SELECTION SWITCH/HARNESS
Agilent 8360 Replaceable P arts 9-47

T able 9-7. Rear P anel (2 of 2)
Ref.
Desig.
16 F AN (see T able 9-2 )
6960-0027 4 PLUG-HOLE TR-HD FOR .625-D-HOLE NYL
18 1 J11-CA A Y-AM/FM OUTPUT (Option 002)
2190-0102 W ASHER LK INTL T 15/32 IN .472-IN-ID
6960-0041 1 PLUG-HOLE TR-HD FOR .5-D-HOLE NYL (Standard)
19 1 J12-CA A Y-PULSE SYNC OUT (Option 002)
2190-0102 W ASHER LK INTL T 15/32 IN .472-IN-ID
6960-0041 1 PLUG-HOLE TR-HD FOR .5-D-HOLE NYL (Standard)
20 1 J13-CA A Y-PULSE VIDEO OUT (Option 002)
2190-0102 W ASHER LK INTL T 15/32 IN .472-IN-ID
6960-0041 1 PLUG-HOLE TR-HD FOR .5-D-HOLE NYL (Standard)
22 08360-20245 1 REAR P ANEL INTERF A CE SHIELD
23 0515-2032 2 SCREW-MA CH M3 x 0.5 8MM-LG 90-DEG-FLH-HD
A T1 1810-0118 1 TERMINA TION SMA MALE 50
F1 2110-0227 1 FUSE (INCH) 5A 250V NTD FE UL
2110-0003 1 FUSE (INCH) 3A 230V NTD FE UL
2110-0565 1 FUSEHOLDER CAP
2110-0569 1 FUSEHOLDER NUT
9-48 Replaceable P arts Agilent 8360

Rear P anel F an Assembly

Figure 9-8. Rear P anel F an Assembly
T able 9-8. Rear P anel F an Assembly
Ref.
Desig.
2 08360-00059 1 F AN SHR OUD
3 08360-20254 1 FO AM
4 08360-00058 1 HEX CELL HONEYCOMB
5 08360-40011 1 GR OUND HEX SHR OUD
6 08360-00060 1 REAR P ANEL
7 0515-2011 8 SCREW-MA CHINE ASSY M3.5 x 0.6
9 08360-00057 2 BRA CKET F AN/REAR P ANEL
11 08360-80026 1 F AN
Agilent 8360 Replaceable P arts 9-49
P art · Qty Description · Num b er

Bottom View, A ttaching Hardw are

Figure 9-9. Bottom View, A ttaching Hardw are
T able 9-9. Bottom View, A ttaching Hardw are
Ref.
Desig.
1 6960-0016 2 PLUG-HOLE TR-HD FOR .125-D HOLE NYL
2 SCREW-MA CHINE (See T able 9-7 )
3 0624-0518 3 SCREW-TPG 6-19 .25-IN-LG P AN-HD-TORX T15
4 08360-40002 1 HIGH V OL T A GE CO VER
5 0515-0382 16 SCREW-MA CHINE ASSEMBL Y M4 x 0.7 12MM-LG
6 SCREW-MA CHINE (See T able 9-4 )
7 SCREW-MA CHINE (See T able 9-4 )
9-50 Replaceable P arts Agilent 8360
P art · Qty Description · Num b er

T op View, A ttaching Hardw are

Figure 9-10. T op View, A ttaching Hardw are
Agilent 8360 Replaceable P arts 9-51

T able 9-10. T op View, A ttaching Hardw are
Ref.
Desig.
2 SCREW-MA CHINE (See T able 9-8 )
4 0515-0501 2 SCREW-SKT-HD-CAP M6 x 1.0 12MM-LG
(83620B/22B/30B/40B/50B)
5 08360-00008 1 MOUNTING BRA CKET COUPLER/DETECTOR (83623B /24B)
6 0515-0372 2 SCREW-MA CHINE ASSEMBL Y M3 x 0.5 8MM-LG (Quan tity is 4 in
instruments with Option 001)
8 0515-0430 2 SMM 3.0 6 PN TX
9 08360-00035 1 SUPPOR T CONNECTORS
10 08360-20237 1 EXTR USION-FIL TER MOUNT
0515-0383 1 SCREW-MA CHINE ASSEMBL Y M4 x 0.7 16MM-LG
12 SCREW-MA CHINE (See T able 9-4 )
13 SCREW-MA CHINE (See T able 9-4 )
15 8160-0539 RFI R OUND STRIP CU-CLD-STL SN-PL (2.56 FT)
16 0515-0382 6 SCREW-MA CHINE ASSEMBL Y M4 x 0.7 12MM-LG
18 08360-20045 1 PO WER SUPPL Y CO VER
19 2200-0103 2 SCREW-MA CH. 4-40 .25-IN-LG P AN-HD-POZI
1520-0205 3 SHOCK MOUNT .31-EFF-HGT
20 0515-1539 4 SCREW-MA CHINE M3 x0.5 16MM-LG
22 08360-00031 1 MOUNTING BRA CKET STEP A TTENUA TOR
9-52 Replaceable P arts Agilent 8360
P art · Qty Description · Num b er

Left Side View, A ttaching Hardw are

Figure 9-11. Left Side View, A ttaching Hardw are
T able 9-11. Left Side View, A ttaching Hardw are
Ref.
Desig.
1 0515-2086 4 SCREW-SPCL M4 x 0.7 7MM-LG 90-DEG-FLH-HD
2 0515-0382 15 SCREW-MA CHINE ASSEMBL Y M4 x 0.7 12MM-LG
3 0515-2086 SCREW-SPCL M4 x 0.7 7MM-LG 90-DEG-FLH-HD
Agilent 8360 Replaceable P arts 9-53
P art · Qty Description · Num b er

Right Side View, A ttaching Hardw are

Figure 9-12. Right Side View, A ttaching Hardw are
T able 9-12. Right Side View, A ttaching Hardw are
Ref.
Desig.
1 0515-0382 33 SCREW-MA CHINE ASSEMBL Y M4 x 0.7 12MM-LG
2 0515-2086 8 SCREW-SPCL M4 x 0.7 7MM-LG 90-DEG-FLH-HD
3 0515-0433 2 SCREW-MA CHINE ASSEMBL Y M4 x 0.7 8MM-LG
4 0515-0382 SCREW-MA CHINE ASSEMBL Y M4 x 0.7 12MM-LG
5 0515-2086 SCREW-SPCL M4 x 0.7 7MM-LG 90-DEG-FLH-HD
9-54 Replaceable P arts Agilent 8360
P art · Qty Description · Num b er

RF Deck Right Side View, A ttaching Hardw are

Figure 9-13. RF Deck Right Side View, A ttaching Hardw are
T able 9-13. RF Deck Right Side View, A ttaching Hardw are
Ref.
Desig.
1 0515-0372 17 SCREW-MA CHINE ASSEMBL Y M3 x 0.5 8MM-LG
2 3050-0169 4 W ASHER-SPR CR VD NO.6 .143-IN-ID
3 0515-2011 4 SCREW-MA CHINE ASSEMBL Y M3.5 x 0.060 12MM-LG
4 2360-0370 4 SCREW-MA CH 6-32 .375-IN-LG P AN-HD-POZI
5 0515-0372 SCREW-MA CHINE ASSEMBL Y M3 x 0.5 8MM-LG
7 0515-0372 SCREW-MA CHINE ASSEMBL Y M3 x 0.5 8MM-LG
8 0515-0458 2 SCREW-MA CHINE ASSEMBL Y M3 x 0.6 8MM-LG (Option 006)
9 0515-0430 2 SMM3.0 6 PN TX (Used for bridge detector regardless of its p osition on
Agilent 8360 Replaceable P arts 9-55

RF Connector and A ttaching Hardw are

Figure 9-14. RF Connector and A ttaching Hardw are
T able 9-14. RF Connector and A ttaching Hardw are
Ref.
Desig.
1 08360-20095 1 FRONT MOUNTING BRA CKET 3.5 MM CONNECTOR
08360-20096 1 FRONT MOUNTING BRA CKET 2.4 MM CONNECTOR
2 08360-20088 1 SP A CER FOR 3.5 MM CONNECTOR
3 2950-0001 1 NUT-HEX-DBL-CHAM 3/8-32-THD (3.5 mm connector)
08360-20133 1 NUT-HEX-DBL-CHAM M10 X1 (2.4 mm connector)
5 08360-00004 1 PL-RR CONN 3.5 (Option 004)
08360-00024 1 PL-RR CONN 2.4 (Option 004)
J1 RF CONNECTOR (See T able 9-2 )
9-56 Replaceable P arts Agilent 8360
P art · Qty Description · Num b er


Figure 9-15. Chassis P arts
Agilent 8360 Replaceable P arts 9-57
Chassis P arts

T able 9-15. Chassis P arts
Ref.
Desig.
2 0515-1245 1 SCREW - CO VER
3 5021-5838 2 SIDE RAILS
4 5041-8802 1 TRIM STRIP
5 5021-8405 1 FRONT FRAME
6 5062-3800 2 HANDLE ASSEMBL Y
7 0515-0943 4 HANDLE SCREWS
8 5021-8497 2 TRIM FRONT HANDLE
9 5041-8801 4 FOOT -FM .5M
10 5062-3748 1 CO VER-BOTTOM
11 1460-1345 2 TIL T ST AND
12 5062-3843 2 CO VER-SIDE
13 5041-8819 2 CAP-FR ONT
14 0515-0707 4 STRAP SCREW
15 5041-8820 2 CAP-REAR
16 5062-3705 2 STRAP HANDLE
17 08360-20040 1 SIDE FRAME
18 5021-5806 1 REAR FRAME
19 54110-40502 4 FOOT REAR
20 0515-1444 4 SCREW - REAR FOOT
21 08360-20253 1 INSULA TED BOTTOM CO VER
9-58 Replaceable P arts Agilent 8360
P art · Qty Description · Num b er

T able 9-16. Miscellaneous Replaceable Accessories
Ref.
Desig.
7120-3737 1 LBL WRN HI V OL T
83620-80004 1 NAMEPLA TE 83620B
83621-80002 1 NAMEPLA TE 83621B
83622-80002 1 NAMEPLA TE 83622B
83623-80003 1 NAMEPLA TE 83623B
83624-80002 1 NAMEPLA TE 83624B
83630-80002 1 NAMEPLA TE 83630B
83631-80002 1 NAMEPLA TE 83631B
83640-80003 1 NAMEPLA TE 83640B
83650-80002 1 NAMEPLA TE 83650B
83651-80002 1 NAMEPLA TE 83651B
83623-80004 1 NAMEPLA TE 83623L
83630-80003 1 NAMEPLA TE 83630L
83640-80004 1 NAMEPLA TE 83640L
83650-80003 1 NAMEPLA TE 83650L
SER VICE TOOLS
T-10 TORX SCREWDRIVER
T-15 TORX SCREWDRIVER
#1 x 3 1/8 IN. POZIDRIVE SCREWDRIVER
#2 x 4 IN. POZIDRIVE SCREWDRIVER
7 MM NUT DRIVER
.050 HEX DRIVER
9/16 IN. NUT DRIVER
5/16 IN OPEN END WRENCH
14MM OPEN END WRENCH
1251-2170 CO AX EXTRA CTOR TOOL-SUBMIN D CONNECTOR
08360-60065 SUBSTITUTE FRONT P ANEL
Agilent 8360 Replaceable P arts 9-59
P art · Qty Description · Num b er

T able 9-16. Miscellaneous Replaceable Accessories (continued)
Ref.
Desig.
08360-60086 EXTENDER BO ARD-DIGIT AL BO ARDS
08360-60085 EXTENDER BO ARD-ANALOG BO ARDS
08360-60081 EXTENDER BO ARD-POST REGULA TOR
08360-60077 EXTENDER BO ARD-SWITCHING REGULA TOR
08360-60030 EXTENDER BO ARD-RECTIFIER/FIL TER
08360-60084 EXTENDER BO ARD-RF DECK
08360-20286 RF TEST CABLE-FIL TER TO DIRECTIONAL COUPLER
08360-60108 EXTENDER CABLE-PO WER SUPPL Y
8710-1787 5 MM HEX BALL DRIVER
ACCESSORIES
TOUCH-UP P AINT
6010-1146 DO VE GRA Y - FOR FRAME AR OUND FRONT P ANEL & P AINTED
6010-1147 FRENCH GRA Y - FOR SIDE, TOP , & BOTTOM CO VERS
6010-1148 P AR CHMENT GRA Y - FOR RA CK MOUNT FLANGES, RA CK
5041-9148 PC BO ARD EXTRA CTORS
5062-7202 CABLE SMI to 8349B
1250-2187 AD APTER 2.4 MM (F) TO K (F)
1250-2188 AD APTER 2.4 MM (F) TO 2.4 MM (F)
5061-5311 AD APTER 3.5 MM (F) TO 3.5 MM (F)
08360-10001 1 DISK - SER VICE SUPPOR T SOFTW ARE
9-60 Replaceable P arts Agilent 8360

T able 9-16. Miscellaneous Replaceable Accessories (continued)
Ref.
Desig.
Do cumen tation F or 8360 B-Series Sw ept Signal Generator
08360-90128 8360 B-Series Man ual Set {
(This includes the following four man uals:)
08360-90126 Quick Reference Guide
08360-90124 T roublesho oting Guide
08360-90179 Obsolete Blo ck Diagrams (not include d in manual set )
Do cumen tation F or 8360 L-Series Sw ept CW Generator
08360-90135 8360 L-Series Man ual Set {
(This includes the following four man uals:)
08360-90126 Quick Reference Guide
08360-90124 T roublesho oting Guide
08360-90179 Obsolete Blo ck Diagrams (not include d in manual set )
Do cumen tation F or 83621B/31B/51B
08360-90138 83621B/31B/51B Man ual Set
(This includes the following man uals:)
08360-90136 T roublesho oting Guide
Agilent T ec hnologies 8360 Comp onen t-Lev el Repair Man ual
08360-90141 Agilent T ec hnologies 8360 Comp onent-Level Repair Man ual
Agilent 8360 Replaceable P arts 9-61

T able 9-16. Miscellaneous Replaceable Accessories (continued)
Ref.
Desig.
08360-60218 ADD OPTION 001 RETR OFIT KIT - 83620B/22B/24B, 83623B/L, and
08360-60220 ADD OPTION 001 RETR OFIT KIT - 83640B/L and 83650B/L
08360-60129 DEL OPTION 001 RETR OFIT KIT - 83620B/22B/24B, 83623B/L, and
08360-60221 DEL OPTION 001 RETR OFIT KIT - 83640B/L and 83650B/L
83602A ADD OPTION 002 RETR OFIT KIT (8360 B- Series)
08360-60236 ADD OPTION 004 RETR OFIT KIT - 83620B/22B/24B, 83623B/L, and
08360-60237 ADD OPTION 004 RETR OFIT KIT - 83640B/L and 83650B/L
83620B/22B/23B/24B/30B
08360-60238 DEL OPTION 004 RETR OFIT KIT -
83640B/50B
08360-60141 ADD OPTION 008 RETR OFIT KIT
08360-60142 ADD OPTION 700 RETR OFIT KIT
08360-60143 ADD OPTION 806 RETR OFIT KIT
08360-60144 ADD OPTION 908 RETR OFIT KIT
08360-60145 ADD OPTION 913 RETR OFIT KIT
08360-60201 FIRMW ARE UPGRADE KIT
9-62 Replaceable P arts Agilent 8360
P art · Qty Description · Num b er

Prev entiv e Maintenance
This c hapter con tains the follo wing information on care and maintenance of the syn thesizer:
Ho w to Clean the F an Filter
Ho w to Clean the Displa y Filter
Ho w to Replace the Line F use
Ho w to Replace the CPU Battery
This is a Safety Class I product (pro vided with a protectiv e earthing ground
incorporated in the po w er cord). The mains plug shall only be inserted in a
sock et outlet pro vided with a protectiv e earth contact. Any interruption of the
protectiv e conductor, inside or outside the instrument, is lik ely to mak e the
instrument dangerous. Intentional interruption is prohibited.
Agilent 8360 Prev entiv e Maintenance 10-1
Introduction · Connector Care

How to Clean the F an Filter
The co oling fan, lo cated on the rear panel, has a thin foam lter. Ho w often the lter m ust b e
cleaned dep ends on the en vironment in whic h the syn thesizer op erates. As the lter collects
dust, the fan sp eed increases to maintain air ow (as the fan sp eed increases, so do es the fan
noise). If the lter con tinues to collect dust after the fan reac hes maximum sp eed, air ow is
reduced and the syn thesizer's in ternal temp erature increases. If the in ternal temp erature
C, the syn thesizer automatically turns o and the amb er standb y LED turns on.
The detachable po w er cord is the instrument disconnecting device. It
disconnects the mains circuits from the mains supply before other parts of the
instrument. The front panel switch is only a standb y switch and is not a LINE
switch.
Clean the fan lter as follows:
10-2 Prev entiv e Maintenance Agilent 8360
reac hes 90 ·


Figure 10-1. Remo ving the F an Filter
Agilent 8360 Prev entiv e Maintenance 10-3

How to Clean the Displa y Filter
The syn thesizer's display is protected b y a plastic display lter. T o clean the display lter, use
mild soap or detergen t and w ater, or a commercial window cleaner (ammonia do es not h urt
the plastic surface). Use a soft, lint-free cloth. Do not use abrasiv e cleaners, tissues, or pap er
tow els whic h can scratch the plastic.
Under normal op eration, y ou will need to clean only the front side of the display lter.
Occasionally , dep ending on the en vironment in whic h the syn thesizer op erates, y ou may wish
to clean the bac k side of the lter as follo ws:
The detachable po w er cord is the instrument disconnecting device. It
disconnects the mains circuits from the mains supply before other parts of the
instrument. The front panel switch is only a standb y switch and is not a LINE
switch.
When y ou remov e the front panel, if y ou do not disconnect the A C line cord
y ou can easily blow the fuse on the preregulator assembly .
and the source mo dule in terface.
10-4 Prev entiv e Maintenance Agilent 8360

How to Replace the Line Fuse
The line fuse is lo cated on the instrumen t rear panel. Replace this fuse as follows:
The detachable po w er cord is the instrument disconnecting device. It
disconnects the mains circuits from the mains supply before other parts of the
instrument. The front panel switch is only a standb y switch and is not a LINE
switch.
coun terclo c kwise, and remov e the fuse holder.
For continued protection against fire hazard replace line fuse only with same
type and rating (refer to Figure 10-2). The use of other fuses or material is
prohibited.
clo c kwise to secure the fuse holder in place.

Figure 10-2. Replacing the Line Fuse
Agilent 8360 Prev entiv e Maintenance 10-5

How to Replace the CPU Battery (A15BT1)
same or equiv alent type recommended.
Battery A15BT1 contains lithium iodide. Do not incinerate or puncture this
battery . Dispose of the discharged battery in a safe manner.
Do not thro w batteries a w ay but collect as small chemical w aste.
The A15 CPU Assem bly may b e damaged if it is placed on a conductiv e
surface. Use a static-safe w ork station. Ensure that the CPU Assem bly is not
placed on an y conductiv e material.
Chapter 7 pro cedure in Chapter 7.

Figure 10-3. A15BT1 Battery Location
Placing the b ottom side of the A15 assembly on a conductiv e w ork surface
without an insulator can result in loss of correction data stored in RAM.
Protect the A15 assembly from con tact with an y conductiv e surface.
on the A15 assembly .
10-6 Prev entiv e Maintenance Agilent 8360
Remo v al/Replacement

Clean and gage the syn thesizer's output connector p erio dically . F or details, refer to
Applic ation Note 326 , whic h do cumen ts the principles of microw av e connector care.
Agilent 8360 Prev entiv e Maintenance 10-7
Connector Care



This c hapter provides option retro t information. Eac h option and the complexity of the
retro t pro cedure are brie y describ ed. Some options cannot b e retro tted and so are not
men tioned in this c hapter. Where applicable, kit part n um b ers are provided for b oth adding
and deleting the option. The sp eci c details of the retro t, including an y adjustmen ts and
p erformance tests required after an assembly is replaced, are discussed in the installation
n um b er.
Option Numb er Option Name
001 Add A tten uator
002 Add Mo dulation/Generator
004 Rear P anel RF Output
008 1 Hz Capabilit y
700 MA TE Compatibilit y
806 Rack Moun t Side Kit
908 Rack Flange Kit Without Handles
910 Extra Man ual Set
913 Rack Flange Kit With Handles
Agilent 8360 Option Retrofits 11-1
Option Retrofits · Introduction · Option Retrofits

Add A ttenuator (Add Option 001)
Option 001 adds an attenuator to the RF dec k to provide the capabilit y of setting a
wider range of output p ow er. T o add the attenuator, the RF dec k m ust b e recon gured.
Adjustmen ts and p erformance tests that are required after installation of the attenuator
are listed in Chapter 8 and in the installation note. Because the step attenuator atness
test requires v ery sp eci c test equipmen t that is not part of the standard microw a v e test
system, and b ecause sev eral exp ensive micro circuits m ust b e handled in the pro cedure, it is
recommended that the instrumen t b e returned to an Agilent service cen ter for retro tting.
Retro t Kit P art Numb er
Mo del Numb er
83622B
83630B/L
83623B/L 08360-60252
83624B
83640B/L 08360-60220
83650B/L
Delete A ttenuator (Delete Option 001)
Retro t Kit P art Numb er
Mo del Numb er
83623B/L
83624B
83630B/L
83640B/L
83650B/L
11-2 Option Retrofits Agilent 8360
83620B None

Add Modulation/Generator (Add Option 002)
Option 002 adds a digitally-synthesized in ternal mo dulation source and AM depth and FM
deviation meters. Because this is a complex pro cedure, it is recommended that the instrumen t
b e returned to an Agilent service cen ter for retro tting. Purc hase of the retro t kit includes
installation at an Agilent service cen ter.
Retro t Kit Mo del Numb er
Mo del Numb er
All B-series mo dels Agilent 83602A
Agilent 8360 Option Retrofits 11-3

Rear P anel RF Output (Add Option 004)
Option 004 mov es the RF output, external ALC input, pulse input/output,
AM input, and FM input connectors from the front panel to the rear panel. Due to the
unique design of the RF dec k, there is no loss of output p ow er with the rear output option.
Adjustmen ts and p erformance tests that are required after installation are listed in Chapter 8
and in the installation note included in the retro t kit.
Retro t Kit P art Numb er
Mo del Numb er
83622B
83623B
83624B
83630B
83640B 08360-60237
83650B
Front P anel RF Output (Delete Option 004)
Retro t Kit P art Numb er
Mo del Numb er
83622B
83623B
83624B
83630B
83640B 08360-60238
83650B
11-4 Option Retrofits Agilent 8360

1 Hz Capability (Add Option 008)
Option 008 provides 1 Hz frequency resolution (1 kHz resolution is standard).
HP BASIC 5.1 and an HP 9000 series 200 or 300 computer are required to run the softw are to
retro t the instrumen t.
Retro t Kit P art Numb er
Mo del Numb er
All mo dels 08360-60141
MA TE Compatibility (Add Option 700)
Option 700 adds CI IL program commands to the instrumen t for MA TE system compatibilit y .
HP BASIC 5.1 and an HP 9000 series 200 or 300 computer are required to run the softw are to
retro t the instrumen t.
Retro t Kit P art Numb er
Mo del Numb er
All mo dels 08360-60142
Agilent 8360 Option Retrofits 11-5

Rack Mount Slide Kit (Add Option 806)
Option 806 con tains the necessary hardw are to moun t sliding rack moun ts on the syn thesizer.
This allo ws easier access to the syn thesizer when it is moun ted in an equipmen t rack. Y ou
m ust remov e the instrumen t side panels to install the kit. Instructions for installation are in
the installation note included in the retro t kit.
Retro t Kit P art Numb er
Mo del Numb er
All mo dels 08360-60143
Rack Flange Kit Without Handles (Add Option 908)
Option 908 con tains a pair of anges and the necessary hardw are to moun t the syn thesizer
( without hand les ) in an equipmen t rack with 482.6 mm (19 in.) horizontal spacing.
Instructions for installation are in the installation note included in the retro t kit.
Retro t Kit P art Numb er
Mo del Numb er
All mo dels 08360-60144
11-6 Option Retrofits Agilent 8360

Extra Manual Set (Add Option 910)
All instrumen ts are supplied with one complete man ual set whic h do cumen ts op eration and
service. Eac h Option 910 (av ailable only with the initial order) provides one additional
complete man ual set. T o order additional man uals after initial shipmen t, order the individual
man ual part n um b ers, or the man ual set part n um b ers listed in Chapter 9.
Rack Flange Kit With Handles (Add Option 913)
Option 913 con tains a pair of anges and the necessary hardw are to moun t the syn thesizer
( with hand les attache d ) in an equipmen t rack with 482.6 mm
(19 in.) horizontal spacing. Instructions for installation are in the installation note included in
the retro t kit.
Retro t Kit P art Numb er
Mo del Numb er
All mo dels 08360-60145
Agilent 8360 Option Retrofits 11-7



How to Use Instrument History
This man ual do cumen ts the curren t pro duction v ersions of the \standalone" Agilent
8360 B-Series Sw ept Signal Generator/ 8360 L-Series Sw ept CW Generator.
This man ual will b e mo di ed to apply to future v ersions of these instrumen t mo dels.
Information provided in this c hapter will then allo w y ou to adapt this man ual to the earlier
v ersions. If y our instrumen t is one of the earlier v ersions, y ou may ha v e to mo dify y our
man ual using the information in this c hapter. Chec k the serial n um b er pre x attac hed to y our
syn thesizer's rear panel and then lo cate it in the following tables. The tables tell y ou whic h
c hanges to make. Incorp orate the c hanges in rev erse alphab etical order.
Serial Numb er Pre x Required Changes
3844A No Change Needed
3722A and b elow A
Serial Numb er Pre x Required Changes
3844A No Change Needed
3722A and b elow A
Agilent T ec hnologies In ternal Use Only

The A4 Fractional-N Assembly has Changed
All Agilent 8360 B-Series and 8360 L-Series instrumen ts with serial pre x n um b ers 3722A and
b elow ha v e a F ractional-N assembly that is di eren t than the one do cumen ted in this man ual.
As a result, y ou m ust apply the c hanges describ ed in this section in order for this man ual to
conform to y our instrumen t.
The part n um b ers listed in T able 12-1 are correct for y our instrumen t. Use these part
n um b ers instead of the part n um b ers ha ving the same reference designators in the
\Replaceable P arts" c hapter. Refer to Figure 12-3 for assembly and cable lo cations.
T able 12-1.
Ref.
Desig.
A4 08360-69010 FRA CTIONAL{N ASSEMBL Y (Rebuilt-Exchange)
W12 08360-60075 1 30{60 MHZ{A4J2/A5J2
W19 08360-60096 1 10 MHZ REAR P ANEL {A7J5/J7
12-2 Instrument History Agilent 8360


Figure 12-3. Major Assembly and Cable Locations
Change A · Instrument History 12-3 · Service

The following adjustmen ts apply to y our instrumen t. Use these adjustmen ts in addition to the
ones describ ed in the \Adjustmen ts" c hapter of this man ual.
Fractional-N V CO Adjustment
No test equipmen t is required for this pro cedure.
This adjustmen t sets the V CO tuning v oltage to 6.0 V for a V CO output of 60 MHz.
b oard.
. Let the instrumen t w arm up for at least one
hour.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SERVICE
display .

Figure 12-4. A4L1 Location
Related P erformance T ests
12-4 Instrument History Agilent 8360

In Case of Difficulty
Swept CW Gener ator T r oublesho oting Guide .
Fractional-N Reference and API Spurs Adjustment
Do not p erform this adjustmen t on a new or rebuilt replacemen t assembly .
The A4 assembly is factory-adjusted, and is not instrumen t dep enden t.
This adjustmen t minimizes three kno wn spurs (125 kHz reference, API1, and API2). A t eac h
spur frequency , the sp ectrum analyzer is phase lo c k ed to the syn thesizer, connected directly to
the A4 fractional-N assembly output, and op erated in zero span. A4R70, R31, and R30 are
adjusted for minimum spur lev el. T able 12-2 summarizes this adjustmen t.
T able 12-2. Summary of Fractional-N Reference & API Spurs Adjustment
P oten tiometer Minimum
125 kHz Reference 44 .125 44 .0 A4R70 75
API 1 44 .005 44 .0 A4R31 70
API 2 44 .0005 44 .0055 A4R30 80
Figure 12-5 . Preset the instrumen ts, and let them w arm up for at least one hour.


Figure 12-5. Fractional-N Reference and API Spur Adjustment Setup
Sp ectrum Analyzer HP/Agilent 8566B
125 kHz Reference Spur Adjustment
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SERVICE
CW
44.125
Adjust the frequency o set to cen ter the signal on the display .
P eak the signal on the display .
12-6 Instrument History Agilent 8360

analyzer.

Figure 12-6. Fractional-N Reference and API Spurs Adjustment Locations
The di erence in lev el b et w een the signal noted in step 4 and the signal in this step should b e
at least 75 dBc.
Change A · Instrument History 12-7 · Service

API 1 Spur Adjustment
44.005
analyzer.
The di erence in lev el b et w een the signal noted in step 7 and the signal in this step
should b e at least 70 dBc.
API 2 Spur Adjustment
44.0005
analyzer.
The di erence in lev el b et w een the signal noted in step 10 and the signal in this step
should b e at least 80 dBc.
Related P erformance T ests
In Case of Difficulty
enough to b e hidden b y phase noise.
A4 fractional-N assembly . Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal
Gener ator/ 8360 L-Series Swept CW Gener ator T r oublesho oting Guide .
12-8 Instrument History Agilent 8360

Use the information in T able 12-3 instead of the equiv alen t F ractional{N information provided
in the \Calibration Constants" c hapter.
T able 12-3.
Assemblies and Corresponding Cal Constant ID Numbers and Revisions
Assembly ID Numb er Revision Part Numb er Description
0 08360-60010 F ractional{N
Use the information in T able 12-4 instead of the equiv alen t F ractional{N information provided
in the \Post Repair" c hapter.
T able 12-4.
Adjustments and P erformance T ests Required
After Repair or Replacement of an Assembly
Assembly Adjustmen t Performance T est
F ractional{N V CO
F ull Self{T est
F requency Switching Time
The adjustmen ts listed are required only after r ep air of the assembly . The r eplac ement assembly
is adjusted at the factory and is not instrument-dep enden t.



AM/FM DA C o set and gain adjustmen t, 3-73
10 MHz reference standard remo v al, 7-29
AM input imp edance adjustmen t, 3-86
AM o set adjustmen t, 3-78
adjustmen t interv al, 3-8
ampli er/ lter adjustmen t, 3-32
10 MHz standard adjustmen t, 3-4
ampli er lter remo v al, 7-44
125 kHz reference spur, 12-5
ampli er multiplier remo v al, 7-44
ampli er remo v al, 7-44
ampli er switc h remo v al, 7-44
A9 pulse b oard jump er adjustmen t, 3-97
ADC adjustmen t, 5-13
10 MHz standard, 3-4
ADC adjustmen t, 5-13
A9 pulse b oard, 3-97
p ow er atness adjustmen t, 5-14
ALC p ow er lev el accuracy, 3-67
step atten uator atness, 5-10
YO delay adjustmen t, 5-12
automated p erformance tests
AM/FM DA C o set and gain, 3-73
step atten uator atness, 5-5
AM input imp edance, 3-86
AM o set, 3-78
ampli er lter, 3-32
external trigger edge selection, 3-94
cop ying disks, 5-3
FM input imp edance, 3-89
soft w are installation, 5-3
fractional-N reference and API spurs, 12-5
A UX OUTPUT disconnection, 7-25
fractional-N V CO, 12-4
low p ow er SRD bias, 3-60
mo dulation generator atness, 3-92
mo dulator o set and gain, 3-62
p ow er atness, 3-71
sampler assem bly, 3-10
selftest patc hes, 3-98
blo c king capacitor remo v al, 7-42
square w av e symmetry, 3-84
sw eep ramp, 3-9
remo v al, 7-25
BNC connectors, Option 004
YO driver +10 V reference, 3-14
remo v al, 7-36
YO driver gain and linearity, 3-15
YO lo op gain, 3-17
Agilent T ec hnologies oces, xi
calibration constan ts, 4-1
ALC p ow er lev el accuracy adjustmen t, 3-67
alphanumeric display remo v al, 7-10
AM accuracy adjustmen t, 3-75
AM delay adjustmen t, 3-80
AM dynamic range, 2-87
calibration constan ts utilit y, 5-20

c hec ksum v eri cation, 4-2
internal pulse accuracy, 2-105
coupler remo v al, 7-44
internal timebase aging rate, 2-7
J
DC blo c king capacitor remo v al, 7-42, 7-44
J10 remo v al, 7-25
dio de switc h remo v al, 7-44
J14 remo v al, 7-25
directional coupler remo v al, 7-44
J1 RF output connector remo v al, 7-13, 7-26
J1 RF output connector remo v al, front panel,
J1 RF output connector remo v al, Option 004,
RF dec k, 7-37
J6 remo v al, 7-25
J7 remo v al, 7-25
display lter remo v al, 7-17
display remo v al, 7-10
k eyb oard reassem bly, 7-18
doubler remo v al, 7-44
k eyb oard remo v al, 7-15
dual mo dulator remo v al, 7-44
line lter remo v al, 7-35
calibration constan ts, 4-2
lowband assem bly remo v al, 7-44
external ALC input remo v al, 7-36
low p ow er SRD bias adjustmen t, 3-60
external lev eling, 2-30
external trigger edge selection adjustmen t, 3-94
maxim um FM deviation, 2-100
maxim um lev eled p ow er, 2-28
micro circuit remo v al, 7-44
mo dulation generator atness adjustmen t, 3-92
mo dulation meter, 2-107
mo dulator o set and gain adjustmen t, 3-62
motherb oard disassem bly, 7-62
fan remo v al, 7-24
motherb oard remo v al, 7-74
lter remo v al, 7-44
op eration v eri cation, 2-4
op eration v eri cation form, 2-4
FM gain adjustmen t, 3-82
Option 004 BNC connectors
FM input imp edance adjustmen t, 3-89
remo v al, 7-36
fractional-N reference and API spurs adjustmen t,
fractional-N V CO adjustmen t, 12-4
frequency switc hing time, 2-18
front panel BNC connectors remo v al, 7-14
front panel disassem bly, 7-4
front panel em ulation, 5-17
front panel pro cessor remo v al, 7-9
front panel remo v al, 7-7, 7-65
line fuse replacemen t, 10-5
AM dynamic range, 2-87
external lev eling, 2-30
fuse housing remo v al, 7-34

rear panel, 7-27 , 7-70
frequency switc hing time, 2-18
rear panel interface, 7-31
internal pulse accuracy, 2-105
RF dec k, 7-73
RF output connector, 7-13, 7-26
maxim um FM deviation, 2-100
maxim um lev eled p ow er, 2-28
source mo dule interface, 7-11
mo dulation meter, 2-107
p ow er accuracy, 2-23
v oltage selector switc h, 7-32
p ow er atness, 2-26
pulse mo dulation on/o ratio, 2-52
pulse mo dulation video feedthrough, 2-71
pulse p erformance, 2-55
pulse p erformance (alternate), 2-60
RF dec k, 7-37
single sideband phase noise, 2-48
spurious signals (harmonics & subharmonics),
RF dec k disassem bly, 7-37
spurious signals (line-related), 2-45
RF dec k remo v al, 7-73
spurious signals (non-harmonic), 2-41
RF output connector remo v al, 7-26
sw ept frequency accuracy, 2-11
RF output connector remo v al, front panel, 7-67
p ow er accuracy, 2-23
RF output connector remo v al, rear panel, 7-69
p ow er atness, 2-26
RPG1 rotary pulse generator remo v al, 7-12
p ow er atness adjustmen t, 3-71, 5-14
p ow er sensor calibration factor, 5-21
p ow er sensor con guration, 5-21
pre x n umb er, x
sampler assem bly adjustmen t, 3-10
pulse mo dulation on/o ratio, 2-52
selftest patc hes adjustmen t, 3-98
pulse mo dulation video feedthrough, 2-71
pulse p erformance, 2-55
serial n umb er, x
pulse p erformance (alternate test), 2-60
RF dec k, 7-39
RAM c hec ksum, 4-2
single sideband phase noise, 2-48
calibration constan ts, 4-2
rear panel BNC connectors remo v al, 7-25
source mo dule interface remo v al, 7-11
rear panel disassem bly, 7-21
spur adjustmen ts, 12-5
rear panel interface remo v al, 7-31
spurious signals (Harmonics & subharmonics),
rear panel remo v al, 7-27, 7-70
spurious signals (line-related), 2-45
k eyb oard, 7-18
spurious signals (non-harmonic), 2-41
square w av e symmetry adjustmen t, 3-84
10 MHz reference standard, 7-29
step atten uator atness
connectors, front panel BNC, 7-14
step atten uator atness p erformance test, 5-5
connectors, Option 004 BNC, 7-36
step atten uator remo v al, 7-44
connectors, rear panel BNC, 7-25
DC blo c king capacitor, 7-42
calibration constan ts, 4-2
sw eep ramp adjustmen t, 3-9
sw ept frequency accuracy, 2-11
front panel, 7-7 , 7-9, 7-65
front panel BNC connectors, 7-14
T1 transformer remo v al, 7-30
k eyb oard, 7-15
timebase oscillator adjustmen t, 3-4
transformer remo v al, 7-30

Y
UVEPR OM
YIG oscillator remo v al, 7-44
calibration constan ts, 4-2
YO delay adjustmen t, 5-12
YO driver +10 V reference adjustmen t, 3-14
YO driver gain and linearity adjustmen t, 3-15
V CO tuning v oltage, 12-4
YO lo op gain adjustmen t, 3-17
v oltage selector switc h remo v al, 7-32
v olts/GHz adjustmen t, 3-95
Index-4