Service Guide

Agilent Technologies 8360 B-Series Swept Signal Generator 8360 L-Series Swept CW Generator

Serial Number Prefixes:

This manual applies to any instrument with the following model number and serial number prefix combinations. You may have to modify this manual so that it applies directly to your instrument version. Refer to the “Instrument History” chapter.

Agilent Technologies 83620B/22B/23B/24B/30B 3844A and Below

Agilent Technologies 83640B/50B 4040A and Below

Agilent Technologies 83623L/30L 3844A and Below

Agilent Technologies 83640L/50L 4040A and Below

Agilent Technologies 83640B — Agilent Technologies 83620B/22B/23B/24B/30B 3844A and Below diagram (page 1)
Fig. 1 · Agilent Technologies 83620B/22B/23B/24B/30B 3844A an…

Part Number: 08360-90125

Printed in USA

October 2000

© Copyright 1997-2000 Agilent Technologies, Inc.

Page 2 of the Agilent Technologies 83640B user manual — Questions or Comments about our Documentation?
Page 2 · click to enlarge
Page 2

Notice

The material contained in this document is provided “as is”, and is subject to being changed, without notice, in future editions.

Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied with regard to this manual and to any of the Agilent products to which it pertains, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or any of the Agilent products to which it pertains. Should Agilent have a written contract with the User and should any of the contract terms conflict with these terms, the contract terms shall control.

Questions or Comments about our Documentation?

We welcome any questions or comments you may have about our documentation. Please send us an E-mail at

[email protected].agilent.com.

Page 3 of the Agilent Technologies 83640B user manual — Printed in USA July 2004
Page 3 · click to enlarge
Page 3

Notice

Hewlett-Packard to Agilent Technologies Transition

This manual may contain references to HP or Hewlett-Packard. Please note that Hewlett- Packard's former test and measurement, semiconductor products and chemical analysis businesses are now part of Agilent Technologies. To reduce potential confusion, the only change to product numbers and names has been in the company name prefix: where a product name/number was HP XXXX the current name/number is now Agilent XXXX. For example, model number HP 8648 is now model number Agilent 8648.

Contacting Agilent Sales and Service Offices

The sales and service contact information in this manual may be out of date. The latest service and contact information for your location can be found on the Web at:

http://www.agilent.com/find/assist

If you do not have access to the Internet, contact your field engineer. In any correspondence or telephone conversation, refer to your instrument by its model number and full serial number.

Agilent Technologies 83640B — http://www.agilent.com/find/assist diagram (page 3)
Fig. 2 · http://www.agilent.com/find/assist

Printed in USA July 2004

Page 4 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 4 · click to enlarge
Page 4

Agilent T ec hnologies certi es that this pro duct met its published sp eci cations at the time

of shipmen t from the factory . Agilent T ec hnologies further certi es that its calibration

measuremen ts are traceable to the United States National Institute of Standards and

T ec hnology , to the exten t allo w ed b y the Institute's calibration facility , and to the calibration

facilities of other In ternational Standards Organization mem b ers.

This Agilent T ec hnologies instrumen t pro duct is w arranted against defects in material and

w orkmanship for a p erio d of one y ear from date of shipmen t. During the w arrant y p erio d,

Agilen T ec hnologies Company will, at its option, either repair or replace pro ducts whic h prov e

to b e defectiv e.

F or w arran ty service or repair, this pro duct m ust b e returned to a service facilit y designated

b y Agilent T ec hnologies. Buy er shall prepa y shipping c harges to Agilent T ec hnologies and

Agilent T ec hnologies shall pa y shipping c harges to return the pro duct to Buy er. Ho w ev er,

Buy er shall pa y all shipping c harges, duties, and taxes for pro ducts returned to Agilent

T ec hnologies from another coun try .

Agilent T ec hnologies w arrants that its softw are and rm w are designated b y Agilent

T ec hnologies for use with an instrumen t will execute its programming instructions when

prop erly installed on that instrumen t. Agilent T ec hnologies do es not w arrant that the

op eration of the instrumen t, or softw are, or rm w are will b e unin terrupted or error-free.

Certification · W arranty

Page 5 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 5 · click to enlarge
Page 5

Limit a tion of W arranty

The foregoing w arrant y shall not apply to defects resulting from improp er or inadequate

maintenance b y Buy er, Buy er-supplied softw are or in terfacing, unauthorized mo di cation or

misuse, op eration outside of the en vironmental sp eci cations for the pro duct, or improp er

site preparation or maintenance.

NO OTHER W ARRANTY IS EXPRESSED OR IMPLIED. A GILENT TECHNOLOGIES

SPECIFICALL Y DISCLAIMS THE IMPLIED W ARRANTIES OF MER CHANT ABILITY

AND FITNESS FOR A P AR TICULAR PURPOSE.

THE REMEDIES PR O VIDED HEREIN ARE BUYER'S SOLE AND EX CLUSIVE

REMEDIES. A GILENT TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY

DIRECT, INDIRECT, SPECIAL, INCIDENT AL, OR CONSEQUENTIAL DAMA GES,

WHETHER BASED ON CONTRA CT, TOR T, OR ANY OTHER LEGAL THEOR Y.

Pr o duct maintenanc e agr e ements and other customer assistanc e agr e ements ar e available for

A gilent T e chnolo gies pr o ducts. F or any assistanc e, c ontact your ne ar est A gilent T e chnolo gies

Sales and Servic e Oc e.

c Cop yrigh t Agilent T ec hnologies 1996, 1997, 1999, 2000

All Righ ts Reserv ed. Repro duction, adaptation, or translation without prior written

p ermission is prohibited, except as allo w ed under the cop yright laws.

1400 F oun taingro v e P arkway , San ta Rosa CA, 95403-1799, USA

Exclusive Remedies · Assistance

Page 6 of the Agilent Technologies 83640B user manual — CA UTION
Page 6 · click to enlarge
Page 6

The follo wing safety notes are used throughout this man ual. F amiliarize y ourself with eac h of

the notes and its meaning b efore op erating this instrumen t.

W ARNING

W arning denotes a hazard. It calls attention to a procedure which, if not

correctly performed or adhered to, could result in injury or loss of life. Do

not proceed bey ond a w arning note until the indicated conditions are fully

understood and met.

CA UTION

correctly p erformed or adhered to, w ould result in damage to or destruction

of the instrumen t. Do not pro ceed b ey ond a caution sign un til the indicated

conditions are fully understo o d and met.

Safety Notes

Page 7 of the Agilent Technologies 83640B user manual — W ARNING
Page 7 · click to enlarge
Page 7

W ARNING

These servicing instructions are for use b y qualified personnel only . T o a v oid

electrical shock, do not perform any servicing unless y ou are qualified to do

so.

The opening of co v ers or remo v al of parts is lik ely to expose dangerous

v oltages. Disconnect the instrument from all v oltage sources while it is being

opened.

The detachable po w er cord is the instrument disconnecting device. It

disconnects the mains circuits from the mains supply before other parts of

the instrument. The front panel switch is only a standb y switch and is not a

LINE switch.

The po w er cord is connected to internal capacitors that may remain liv e for 5

seconds after disconnecting the plug from its po w er supply .

This is a Safety Class I product (pro vided with a protectiv e earthing ground

incorporated in the po w er cord). The mains plug shall only be inserted in a

sock et outlet pro vided with a protectiv e earth contact. Any interruption of the

protectiv e conductor, inside or outside the instrument, is lik ely to mak e the

instrument dangerous. Intentional interruption is prohibited.

For continued protection against fire hazard replace line fuse only with same

type and rating (F 5A/250V). The use of other fuses or material is prohibited.

General Safety Considerations

Page 8 of the Agilent Technologies 83640B user manual — CA UTION
Page 8 · click to enlarge
Page 8

CA UTION

Always use the three-prong ac p ow er cord supplied with this instrumen t.

F ailure to ensure adequate earth grounding b y not using this cord may cause

instrumen t damage.

Before switching on this pro duct, make sure that the line v oltage selector

switch is set to the v oltage of the p ow er supply and the correct fuse is

installed. Assure the supply v oltage is in the sp eci ed range.

This pro duct is designed for use in Installation Category I I and P ollution

Degree 2 p er IEC 1010 and 664 resp ectively .

V en tilation Requiremen tsWhen installing the pro duct in a cabinet, the

con v ection in to and out of the pro duct m ust not b e restricted. The ambien t

temp erature (outside the cabinet) m ust b e less than the maximum op erating

temp erature of the pro duct b y 4

C for ev ery 100 w atts dissipated in the

cabinet. If the total p ow er dissipated in the cabinet is greater than 800

w atts, then forced con v ection m ust b e used.

Before Applying P o w er

V erify that the pro duct is con gured to match the av ailable main p ow er source p er the input

p ow er con guration instructions provided in this man ual.

If this pro duct is to b e energized via an autotransformer make sure the common terminal is

connected to the neutral (grounded side of the mains supply).

Page 9 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 9 · click to enlarge
Page 9
Page 9 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 10 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNN
Page 10 · click to enlarge
Page 10

This man ual provides service information for the Agilent T ec hnologies 8360 B-Series swept

signal generator and Agilent T ec hnologies 8360 L-Series swept CW generator.

This represen ts a k ey ph ysically lo cated on the instrumen t.

NNNNNNNNNNNNNNNNNNNNNNN

This indicates a \softkey ," a k ey whose lab el is determined b y the rm w are of

the instrumen t.

Screen Text This indicates text display ed on the instrumen t's screen.

Show 6 more labels / short text on this page
Preface · Manual Conv entions · · F ront-Panel Key · · Softkey
Page 11 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 11 · click to enlarge
Page 11

Instruments Cov ered By This Manual

This man ual applies to instrumen ts ha ving a serial n um b er pre x listed on the title page

(b ehind the \Do cumen tation Map" tab). Some c hanges may ha v e to b e made to this man ual

so that it applies directly to eac h instrumen t; refer to Chapter 12, \Instrumen t History ," to

see what c hanges may apply to y our instrumen t.

A serial n um b er lab el ( Figure 0-1 ) is attached to the instrumen t's rear panel. A pre x (four

digits follo w ed b y a letter), and a sux ( v e digits unique to eac h instrumen t), comprise the

instrumen t serial n um b er.

Agilent Technologies 83640B — 08360125.pdf diagram (page 11)
Fig. 3 · 08360125.pdf

Figure 0-1. Typical Serial Number Label

This man ual applies to the Agilent T ec hnologies 8360 B-Series swept signal generator and

Agilent T ec hnologies 8360 L-Series swept CW generator ha ving a mo del n um b er and serial

n um b er pre x listed on the title page (b ehind the \Do cumen tation Map" tab). Some c hanges

may ha v e to b e made to this man ual so that it applies directly to eac h instrumen t; refer to

Chapter 12, \Instrumen t History" in this man ual to see what c hanges may apply to y our

instrumen t.

This instrumen t has b een designed and tested in accordance with IEC Publication 1010,

Safety Requiremen ts for Electronic Measuring Apparatus, and has b een supplied in a safe

condition. The instruction do cumen tation con tains information and w arnings whic h m ust

b e follo w ed b y the user to ensure safe op eration and to maintain the instrumen t in a safe

condition.

Compliance

Page 12 of the Agilent Technologies 83640B user manual — INTERCON FIELD OPERA TIONS
Page 12 · click to enlarge
Page 12

Sales and Service Offices

Hewlett-P ackard Sales and Service Offices

UNITED ST A TES

Instrument Supp ort Center

EUR OPEAN FIELD OPERA TIONS

Headquarters F rance German y

Hewlett-Pac k ard S.A. Hewlett-Pac k ard F rance Hewlett-Pac k ard Gm bH

150, Route du Nan t-d'Avril 1 Av enue Du Canada Hewlett-Pac k ard Strasse

1217 Meyrin 2/Genev a Zone D'Activite De Courtab o euf 61352 Bad Hom burg v.d.H

Switzerland F-91947 Les Ulis Cedex German y

(41 22) 780.8111 F rance (49 6172) 16-0

(33 1) 69 82 60 60

Hewlett-Pac k ard Ltd.

Eskdale Road, Winnersh T riangle

W okingham, Berkshire R G41 5DZ

INTERCON FIELD OPERA TIONS

Hewlett-Pac k ard Company Hewlett-Pac k ard Australia Ltd. Hewlett-Pac k ard (Canada) Ltd.

3495 Deer Creek Road 31-41 Joseph Street 17500 South Service Road

P alo Alto, California, USA Blackburn, Victoria 3130 T rans- Canada High wa y

94304-1316 (61 3) 895-2895 Kirkland, Queb ec H9J 2X8 Dian District

Japan Singap ore T aiwan

Hewlett-Pac k ard Japan, Ltd. Hewlett-Pac k ard Singap ore (Pte.) Ltd. Hewlett-Pac k ard T aiwan

9-1 T ak akura-Cho, Hac hio ji 150 Beach Road 8th Flo or, H-P Building

T oky o 192, Japan #29-00 Gatew a y W est 337 F u Hsing North Road

(81 426) 60-2111 Singap ore 0718 T aip ei, T aiwan

(65) 291-9088 (886 2) 712-0404

China Hewlett-Pac k ard Company

38 Bei San Huan X1 Road

Shuang Y u Shu

Show 12 more labels / short text on this page
Agilent T echnologies · (800) 403-0801 · Great Britain · England · (44 734) 696622 · Headquarters Australia Canada · (415) 857-5027 Canada · (514) 697-4232 · China · Hai Dian District · Beijing, China · (86 1) 256-6888
Page 13 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 13 · click to enlarge
Page 13
Page 13 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 15 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 15 · click to enlarge
Page 15

In Case of Diculty 2-22

4. P ow er Accuracy 2-23

Description and Pro cedure 2-23

Related Adjustmen ts 2-24

In Case of Diculty 2-25

5. P ow er Flatness 2-26

Description and Pro cedure 2-26

Lo w Band P ow er Flatness 2-26

High Band P ow er Flatness 2-27

Millimeter Band P ow er Flatness > 20 GHz and 40 GHz 2-27

Millimeter Band P ow er Flatness > 40 GHz 2-27

Related Adjustmen ts 2-27

In Case of Diculty 2-27

6. Maxim um Lev eled P ow er 2-28

Description and Pro cedure 2-28

Related Adjustmen ts 2-29

In Case of Diculty 2-29

7. External Lev eling 2-30

Description and Pro cedure 2-30

Related Adjustmen ts 2-31

In Case of Diculty 2-31

8. Spurious Signals (Harmonics & Subharmonics) 2-32

Description and Pro cedure 2-32

Harmonic Measuremen t < 20 GHz 2-33

Harmonic Measuremen t 20 to 26.5 GHz 2-35

Harmonic Measuremen t 26.5 to 40 GHz 2-37

Harmonic Measuremen t 38 to 50 GHz 2-37

Harmonic/Subharmonic V eri cation Pro cedure < 20 GHz 2-38

Harmonic/Subharmonic V eri cation Pro cedure > 20 GHz 2-39

Related Adjustmen ts 2-40

In Case of Diculty 2-40

9. Spurious Signals (Non-Harmonic) 2-41

Description and Pro cedure 2-41

Fixed Spurious Signals 2-41

Lo w Band O set Spurious Signals 2-43

Lo w Band Mixer Spurs 2-44

Related Adjustmen ts 2-44

In Case of Diculty 2-44

10. Spurious Signals (Line-Related) 2-45

Description and Pro cedure 2-45

Related Adjustmen ts 2-47

In Case of Diculty 2-47

11. Single Sideband Phase Noise 2-48

Description and Pro cedure 2-48

Related Adjustmen ts 2-51

In Case of Diculty 2-51

12. Pulse Mo dulation On/O Ratio (F or B-Series Only) 2-52

Description and Pro cedure 2-52

Related Adjustmen ts 2-54

In Case of Diculty 2-54

13. Pulse P erformance (F or B-Series Only) 2-55

Contents-2 Agilent 8360 · Service

Page 16 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 16 · click to enlarge
Page 16

Description and Pro cedure 2-55

Rise and F all Times 2-55

Pulse Lev eling Accuracy 2-58

Related Adjustmen ts 2-59

In Case of Dicult y 2-59

14. Pulse P erformance (Alternate Pro cedure) (F or B-Series Only) 2-60

Description and Pro cedure 2-60

Rise and F all Times < 20 GHz 2-60

Pulse Lev eling Accuracy < 20 GHz 2-64

Rise and F all Times > 20 GHz 2-67

Pro cedure > 20 to 26.5 GHz 2-68

Pro cedure > 26.5 to 40 GHz 2-68

Pro cedure > 40 GHz to 50 GHz 2-68

Pulse Lev eling Accuracy > 20 GHz 2-68

Pro cedure > 20 to 26.5 GHz 2-69

Pro cedure > 26.5 to 40 GHz 2-69

Pro cedure > 40 to 50 GHz 2-69

Related Adjustmen ts 2-70

In Case of Dicult y 2-70

15. Pulse Mo dulation Video F eedthrough (F or B-Series Only) 2-71

Description and Pro cedure 2-71

System Calibration 2-73

Video Gain 2-73

Lo w Band Video F eedthrough 2-74

Video F eedthrough > 2 GHz 2-77

Related Adjustmen ts 2-77

In Case of Dicult y 2-77

16. AM Accuracy (F or B-Series Only) 2-78

Description and Pro cedure 2-78

Lo w Band < 2 GHz 2-78

F or Instrumen ts with Option 002 only 2-79

High Band > 2 GHz and < 20 GHz 2-80

F or Instrumen ts with Option 002 only 2-81

High Band > 20 GHz { Doubler Engaged 2-81

F or Instrumen ts with Option 002 only: 2-83

Related Adjustmen ts 2-83

In Case of Dicult y 2-83

17. AM Bandwidth (F or B-Series Only) 2-84

Description and Pro cedure 2-84

Related Adjustmen ts 2-86

In Case of Dicult y 2-86

18. AM Dynamic Range (F or B-Series Only) 2-87

Description and Pro cedure 2-87

Normal AM Dynamic Range 2-88

Deep AM Dynamic Range 2-89

Related Adjustmen ts 2-91

In Case of Dicult y 2-91

19. FM Accuracy (F or B-Series Only) 2-92

Description and Pro cedure 2-92

External FM 2-92

In ternal FM (Option 002 only) 2-94

Contents-3 · Service

Page 17 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 17 · click to enlarge
Page 17

Related Adjustmen ts 2-94

In case of Diculty 2-94

20. FM Bandwidth (F or B-Series Only) 2-95

Description and Pro cedure 2-95

Find Quadrature 2-98

DC FM Flatness (Unlo c k ed) 2-99

Related Adjustmen ts 2-99

In Case of Diculty 2-99

21. Maximum FM Deviation (F or B-Series Only) 2-100

Description and Pro cedure 2-100

DC FM ( 100 Hz) (Using HP/Agilen t 8111A or HP/Agilent 8116A) . . 2-100

A C FM (at 100 kHz) (Using HP/Agilent 3325A) 2-102

Related Adjustmen ts 2-104

In Case of Diculty 2-104

22. In ternal Pulse Accuracy (Option 002) (F or B-Series Only) 2-105

Description and Pro cedure 2-105

Related Adjustmen ts 2-106

In Case of Diculty 2-106

23. Mo dulation Meter (Option 002) (F or B-Series Only) 2-107

Description and Pro cedure 2-107

Related Adjustmen ts 2-108

In Case of Diculty 2-108

3. Adjustmen ts

Ho w T o Use This Chapter 3-1

Men u Maps 3-1

Equipmen t Required 3-1

List of Adjustmen ts 3-3

1. 10 MHz Standard 3-4

Description and Pro cedure 3-4

Related P erformance T ests 3-7

In Case of Diculty 3-7

Accuracy V ersus Adjustmen t In terv al 3-8

An Alternate Metho d 3-8

2. Sw eep Ramp 3-9

Description and Pro cedure 3-9

Related P erformance T ests 3-9

In Case of Diculty 3-9

3. Sampler Assem bly 3-10

Description and Pro cedure 3-10

200 MHz Lo op V CO T une Chec k 3-11

Sampler Match Adjustmen t 3-11

200 MHz Lo op Gain Adjustmen t 3-12

IF Gain Adjustmen t 3-12

Related P erformance T ests 3-12

In Case of Diculty 3-12

4. YO Driver +10 V Reference 3-14

Description and Pro cedure 3-14

Related P erformance T ests 3-14

In Case of Diculty 3-14

Contents-4 Agilent 8360 · Service

Page 18 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 18 · click to enlarge
Page 18
5. YO Driver Gain and Linearity 3-15

Description and Pro cedure 3-15

Related P erformance T ests 3-16

In Case of Dicult y 3-16

6. YO Lo op Gain 3-17

Description and Pro cedure 3-17

Related P erformance T ests 3-17

In Case of Dicult y 3-17

7. Ampli er/Multiplier Adjustmen ts 3-18

Description 3-18

Adjustmen t Help 3-19

O set and Gain Adjustmen ts 3-19

Adjusting to the Cen ter of the P assband 3-19

SRD Bias Adjustmen t 3-19

Single Band Delay Comp ensation Adjustmen ts 3-19

Setting the F astest Sw eep Time 3-19

Sw eep Sp eed Related Adjustmen ts 3-19

YTM Bandcross Delay T erms 3-20

General T ec hniques 3-20

Pro cedure 3-21

YTM Gain Adjustmen t 3-25

Band 1 Adjustmen t 3-25

Band 2 Adjustmen t 3-26

Band 3 Adjustmen t 3-26

Band 4 Adjustmen t ( 83630B /L, 83640B /L, and 83650B /L only) . . . 3-27

Band 5 Adjustmen t ( 83640B /L and 83650B /L only) 3-28

Band 6 Adjustmen t ( 83640B /L and 83650B /L only) 3-29

Band 7 Adjustmen t ( 83650B /L only) 3-29

Multi-band Ampli er/Multiplier Delay Comp ensation 3-29

Band 4 Multi-band Ampli er/Multiplier Delay Comp ensation ( 83630B /L,

83640B /L, and 83650B /L only) 3-30

V eri cation 3-31

Related P erformance T ests 3-31

In Case Of Diculty 3-31

8. Ampli er/Filter Adjustmen ts 3-32

Description 3-32

Adjustmen t Help 3-33

O set and Gain Adjustmen ts 3-33

Squegg Clamp Calibration Constants 3-33

Adjusting for the Cen ter of the P assband 3-33

Single Band Delay Comp ensation Adjustmen ts 3-34

Setting the F astest Sw eep Time 3-34

Sw eep Sp eed Related Adjustmen ts 3-34

YTF Bandcross Delay T erms 3-34

Pro cedure 3-37

Ampli er/Filter O set and Gain Adjustmen t 3-41

Band 1 Adjustmen t 3-41

Band 2 Adjustmen t 3-43

Band 3 Adjustmen t 3-44

Band 4 Adjustmen t ( 83630B /L, 83640B /L, and 83650B /L only) . . . 3-46

Band 5 Adjustmen t ( 83640B /L and 83650B /L only) 3-47

Contents-5 · Service

Page 19 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 19 · click to enlarge
Page 19

Band 6 Adjustmen t ( 83640B /L and 83650B /L only) 3-48

Band 7 Adjustmen t ( 83650B /L only) 3-48

Auto T racking V eri cation 3-49

Single-Band Ampli er/Filter Delay 3-51

Band 1 Adjustmen t 3-52

Band 2 Adjustmen t 3-52

Band 3 Adjustmen t 3-53

Band 4 Adjustmen t ( 83630B /L, 83640B /L, and 83650B /L) 3-54

Band 5 Adjustmen t ( 83640B /L and 83650B /L only) 3-54

Band 6 Adjustmen t ( 83640B /L and 83650B /L only) 3-55

Band 7 Adjustmen t ( 83650B /L only) 3-55

Multi-band YTF Delay 3-56

Band 4 Multi-band YTF Delay ( 83630B /L, 83640B /L, and 83650B /L

only) 3-58

Op eration Chec k 3-59

Related P erformance T ests 3-59

In Case Of Diculty 3-59

9. Lo w P ow er SRD Bias 3-60

Description and Pro cedure 3-60

Related P erformance T ests 3-61

In Case of Diculty 3-61

10. Mo dulator O set and Gain 3-62

Description and Pro cedure 3-62

Related P erformance T ests 3-66

In Case of Diculty 3-66

11. ALC P ow er Lev el Accuracy 3-67

Description and Pro cedure 3-67

Lo w Band ALC P ow er Accuracy Adjustmen t 3-68

High Band P ow er Accuracy Adjustmen t 3-69

Related P erformance T ests 3-70

In Case of Diculty 3-70

12. P ow er Flatness 3-71

Description and Pro cedure 3-71

Lo w Band P ow er Flatness 3-71

High Band P ow er Flatness 3-72

Millimeter Band P ow er Flatness 3-72

Related P erformance T ests 3-72

In Case of Diculty 3-72

13. AM/FM DA C O set and Gain (Option 002) (8360 B-Series Only) 3-73

Description and Pro cedure 3-73

Related P erformance T ests 3-74

In Case of Diculty 3-74

14. AM Accuracy (8360 B-Series Only) 3-75

Description and Pro cedure 3-75

AM O set 3-76

Log AM Accuracy 3-76

Linear AM Accuracy 3-77

Related P erformance T ests 3-77

In Case of Diculty 3-77

15. AM O set (8360 L-Series Only) 3-78

Description and Pro cedure 3-78

Contents-6 Agilent 8360 · Service

Page 20 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 20 · click to enlarge
Page 20
16. AM Delay (8360 B-Series Only) 3-80

Description and Pro cedure 3-80

Related P erformance T ests 3-81

In Case of Dicult y 3-81

17. FM Gain (8360 B-Series Only) 3-82

Description and Pro cedure 3-82

Related P erformance T ests 3-83

In Case of Dicult y 3-83

18. Square W av e Symmetry 3-84

Description and Pro cedure 3-84

High Band Adjustmen t 3-84

Lo w Band Adjustmen t 3-85

Related P erformance T ests 3-85

In Case of Dicult y 3-85

19. AM Input Imp edance (8360 B-Series Only) 3-86

Description and Pro cedure 3-86

A. A10 Jump er 3-87

B. A8 Jump er 3-88

Related P erformance T ests 3-88

In Case of Dicult y 3-88

20. FM Input Imp edance (8360 B-Series Only) 3-89

Description and Pro cedure 3-89

A. A11 Jump er 3-90

B. A8 Jump er 3-91

Related P erformance T ests 3-91

In Case of Dicult y 3-91

21. Mo dulation Generator Flatness (Option 002) (8360 B-Series Only) . . . 3-92

Description and Pro cedure 3-92

Related P erformance T ests 3-93

In Case of Dicult y 3-93

22. External T rigger Edge Selection 3-94

Description and Pro cedure 3-94

T riggering on the Leading Edge 3-94

T riggering on the T railing Edge 3-94

23. V olts/GHz 3-95

Description and Pro cedure 3-95

Related P erformance T ests 3-95

In Case of Dicult y 3-96

24. A9 Pulse Board Jump er (W1) 3-97

Description and Pro cedure 3-97

25. Selftest P atches (8360 L-Series Only) 3-98

Description and Pro cedure 3-98

Contents-7 · Service

Page 21 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 21 · click to enlarge
Page 21
4. Calibration Constan ts

In tro duction 4-1

De nition 4-1

Memory Areas 4-2

W orking Data 4-2

Protected Data 4-2

Default Data 4-2

Chec ksum V eri cation 4-2

Calibration Constant P assw ord 4-3

En tering a P assword 4-3

Disabling a P assword 4-3

Setting a P assword 4-4

Bypassing the P assword 4-5

Changing W orking Data Calibration Constants 4-6

Sa ving W orking Data Calibration Constants 4-6

Loading Protected Data Calibration Constants 4-7

Loading Default Data Calibration Constants 4-7

Calibration Constant Descriptions 4-8

5. Automated T ests

In tro duction 5-1

Setting Up the System 5-2

Hardware Requiremen ts 5-2

Op erating System Requiremen ts 5-2

Installing the Softw are 5-3

Disk Files 5-3

Make W orking Copies 5-3

Installing the Program on to a Hard Disk 5-3

Op erating from a Floppy Disk 5-3

Running the Softw are 5-4

Con guration Limitations 5-4

Loading the Softw are 5-4

A UTOMA TED PERF ORMANCE TESTS 5-5

1. Step A tten uator Flatness T est 5-5

Description 5-5

Pro cedure 5-6

2. P ow er Flatness and Accuracy T est 5-7

Description 5-7

Pro cedure 5-7

A UTOMA TED ADJUSTMENTS 5-10

1. Step A tten uator Flatness Adjustmen t 5-10

Description 5-10

Pro cedure 5-10

2. YO Delay Adjustmen t 5-12

Description 5-12

3. ADC Adjustmen t 5-13

Description 5-13

4. P ow er Flatness Adjustmen t 5-14

Description 5-14

Pro cedure 5-14

Utilities 5-17

Contents-8 Agilent 8360 · Service

Page 22 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 22 · click to enlarge
Page 22
1. F ront P anel Em ulation 5-17

Description 5-17

Pro cedure 5-17

2. Calibration Constants 5-20

Description 5-20

Pro cedure 5-20

3. Self-T est 5-20

Description 5-20

4. P ow er Sensor Con guration and Calibration F actor File 5-21

Description 5-21

Loading Sensor Calibration F actors 5-21

T o Load Sensor Cal F actors: 5-21

Men u Ov erview 5-24

6. Men u Maps

ALC Men u 6-3

F requency Men u 6-5

Marker Men u 6-7

Mo dulation Men u 6-9

P ow er Men u 6-11

Service Men u 6-13

Sw eep Men u 6-15

System Men u 6-17

User Cal Men u 6-19

7. Disassembly and Replacemen t Pro cedures

In tro duction 7-1

T o ols Required 7-3

F ront P anel Disassembly and Reassem bly 7-4

T o ols Required 7-4

Description and Pro cedure 7-4

1. Preliminary Steps 7-6
2. F ront P anel Remo v al 7-7
3. A3 F ront P anel Pro cessor Remo v al 7-9
4. A21 Alphan umeric Display Remo v al 7-10
5. A2 Source Mo dule In terface Remo v al 7-11
6. RPG1 Rotary Pulse Generator Remo v al 7-12
7. J1 RF Output Connector Remo v al 7-13
8. F ront P anel BNC Connectors Remo v al 7-14
9. A1 Keyb oard Remo v al 7-15
10. Display Filter Remo v al 7-17
11. A1 Keyb oard Reassem bly 7-18
12. F ront P anel Reassem blies 7-20

Rear P anel Disassembly and Reassem bly 7-21

T o ols Required 7-21

Description and Pro cedure 7-21

1. Preliminary Steps 7-23
2. B1 F an Remo v al 7-24
3. Rear P anel BNC Connectors Remo v al and Aux Output Disconnection . . 7-25
4. J1 RF Output Connector Remo v al (Option 004) 7-26
5. Rear P anel Remo v al 7-27

Contents-9 · Service

Page 23 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 23 · click to enlarge
Page 23
6. A23 10 MHz Reference Standard Remo v al 7-29
7. T1 T ransformer Remo v al 7-30
8. A19 Rear P anel In terface Remo v al 7-31
9. V oltage Selector Switch Remo v al 7-32
10. F use Housing Remo v al 7-34
11. FL1 Line Filter Remo v al 7-35
12. Option 004 BNC Connectors Remo v al 7-36

RF Dec k Disassem bly and Reassem bly 7-37

T o ols Required 7-37

Description and Pro cedure 7-37

1. Preliminary Steps 7-39
2. Lift Up the RF Dec k 7-39
3. A20 RF In terface Remo v al 7-43
4. Micro circuit Remo v al 7-44

Motherb oard Disassembly and Reassem bly 7-62

T o ols Required 7-62

Description and Pro cedure 7-62

1. Preliminary Steps 7-64
2. F ront P anel Remo v al 7-65
3. J1 RF Output Connector Remo v al (F ront P anel) 7-67
4. J1 RF Output Connector Remo v al (Option 004) 7-69
5. Rear P anel Remo v al 7-70
6. Ma jor Assem blies Remo v al 7-72
7. RF Dec k Remo v al 7-73
8. Motherb oard Remo v al 7-74
8. P ost{Repair
9. Replaceable P arts

In tro duction 9-1

Mo dule{Exc hange Program 9-1

Replacemen t Information 9-1

Ordering Information 9-2

Ho w to Order P arts . . . F ast! 9-2

10. Prev en tive Main tenance

In tro duction 10-1

Ho w to Clean the F an Filter 10-2

Ho w to Clean the Displa y Filter 10-4

Ho w to Replace the Line F use 10-5

Ho w to Replace the CPU Battery (A15BT1) 10-6

Remo v al/Replacemen t 10-6

Connector Care 10-7

Contents-10 Agilent 8360 · Service

Page 24 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 24 · click to enlarge
Page 24
11. Option Retro ts

In tro duction 11-1

Add A tten uator (Add Option 001) 11-2

Delete A tten uator (Delete Option 001) 11-2

Add Mo dulation/Generator (Add Option 002) 11-3

Rear P anel RF Output (Add Option 004) 11-4

F ront P anel RF Output (Delete Option 004) 11-4

1 Hz Capability (Add Option 008) 11-5

MA TE Compatibilit y (Add Option 700) 11-5

Rac k Moun t Slide Kit (Add Option 806) 11-6

Rac k Flange Kit Without Handles (Add Option 908) 11-6

Extra Manual Set (Add Option 910) 11-7

Rac k Flange Kit With Handles (Add Option 913) 11-7

12. Instrumen t History

Ho w to Use Instrumen t History 12-1

Change A 12-2

The A4 F ractional-N Assem bly has Changed 12-2

Replaceable P arts 12-2

Adjustmen ts 12-4

F ractional-N V CO Adjustmen t 12-4

Description and Pro cedure 12-4

Related P erformance T ests 12-4

In Case of Dicult y 12-5

F ractional-N Reference and API Spurs Adjustmen t 12-5

Description and Pro cedure 12-5

125 kHz Reference Spur Adjustmen t 12-6

API 1 Spur Adjustmen t 12-8

API 2 Spur Adjustmen t 12-8

Related P erformance T ests 12-8

In Case of Dicult y 12-8

Calibration Constants 12-9

P ost Repair 12-9

Index · Contents-11 · Service

Page 25 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 25 · click to enlarge
Page 25

0-1. T ypical Serial Num b er Lab el x

2-1. In ternal Timebase: Aging Rate T est Setup 2-8

2-2. Sw ept F requency Accuracy T est Setup 2-11

2-3. Video Signal on the Oscilloscop e 2-15

2-4. F requency Switching Time T est Setup 2-18

2-5. P ow er Accuracy T est Setup 2-23

2-6. P ow er Flatness T est Setup 2-26

2-7. Maximum Lev eled P ow er T est Setup 2-28

2-8. External Lev eling T est Setup 2-30

2-9. Spurious Signals (Harmonic) T est Setup < 20 GHz 2-32

2-10. Spurious Signals (Harmonics) T est Setup > 20 GHz 2-35

2-11. Spurious Signals (Non-Harmonic) T est Setup 2-41

2-12. Spurious signals (Line-Related) T est Setup 2-45

2-13. Single Sideband Phase Noise T est Setup 2-49

2-14. Pulse Mo dulation On/O Ratio T est Setup 2-52

2-15. Pulse P erformance T est Setup 2-55

2-16. Pulse En v elop e Display ed Ov er 5 Divisions 2-57

2-17. Widened Pulse En v elop e 2-58

2-18. Pulse P erformance T est Setup < 20 GHz 2-61

2-19. Signal Lev el b et w een RF P ow er O and On 2-64

2-20. Pulse P erformance T est Setup >20 GHz 2-67

2-21. Video F eedthrough T est Setup 2-71

2-22. Measuring Maximum V oltage from the Baseline Video F eedthrough 2-76

2-23. AM Accuracy T est Setup Lo w Band 2-78

2-24. AM Accuracy T est Setup < 20 GHz 2-80

2-25. AM Accuracy T est Setup > 20 GHz 2-82

2-26. AM Bandwidth T est Setup 2-84

2-27. AM Dynamic Range T est Setup 2-87

2-28. FM Accuracy T est Setup 2-92

2-29. FM Bandwidth T est Setup 2-97

2-30. Maximum FM Deviation T est Setup 2-100

2-31. Pulse Accuracy T est Setup 2-105

2-32. Mo dulation Meter T est Setup 2-107

3-1. 10 MHz Standard Adjustmen t Setup 3-5

3-2. 10 MHz Standard Adjustmen t Lo cation 3-6

3-3. Accuracy V ersus Adjustmen t In terv al 3-8

3-4. FL2 Lo cation 3-10

3-5. Sampler Match Adjustmen t Lo cations 3-11

3-6. IF Gain Adjustmen t Lo cation 3-12

3-7. YO Driv er +10 V Reference Adjustmen t Lo cation 3-14

3-8. YO Gain and Linearity Break P oints 3-15

3-9. YO Gain and Linearity Adjustmen t Lo cations 3-16

Figures · Contents-12 Agilent 8360 · Service

Page 26 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 26 · click to enlarge
Page 26

3-10. Switch and Adjustmen t Lo cations 3-17

3-11. Ampli er/Multiplier Adjustmen ts Setup 3-21

3-12. A12 Assem bly P oten tiometer Lo cations 3-27

3-16. Single-Band Delay and Risetime Comp ensation 3-35

3-17. Multi-Band Delay and Risetime Comp ensation 3-36

3-18. Ampli er/Filter Adjustmen ts Setup 3-37

3-19. XA10J2 Lo cation 3-38

3-20. A12 Assem bly P oten tiometer Lo cations 3-45

3-21. Lo w P ow er SRD Bias Adjustmen t Setup 3-60

3-22. A9R105 Lo cation 3-61

3-23. Mo dulator O set and Gain Adjustmen t Setup 3-62

3-24. ALC P ow er Lev el Accuracy Adjustmen t Setup 3-67

3-25. A10R125 Lo cation 3-69

3-26. P ow er Flatness Adjustmen t Setup 3-71

3-27. A8R111 and A8R105 Adjustmen t Lo cations 3-73

3-28. AM Accuracy Adjustmen t Setup 3-75

3-29. A10R35 and A10R39 Lo cations 3-77

3-30. AM O set Adjustmen t Setup 3-79

3-31. AM Delay Adjustmen t Setup 3-80

3-32. AM Delay Adjustmen t Lo cation 3-81

3-33. FM Gain Adjustmen t Setup 3-82

3-34. FM Gain Adjustmen t Lo cation 3-83

3-35. Square W av e Symmetry Adjustmen t Setup 3-84

3-36. A10 Jump er Lo cations 3-87

3-37. A8 Jump er Lo cations 3-88

3-38. FM Input Imp edance Jump er Lo cations 3-90

3-39. A8 Jump er Lo cations 3-91

3-40. Mo dulation Generator Flatness Adjustmen t Setup 3-92

3-41. External T rigger Edge Selection Lo cation 3-94

3-42. A12S1 Lo cation 3-95

5-1. Step A tten uator Flatness T est Setup 5-6

5-2. Step A tten uator Flatness Adjustmen t Setup 5-11

6-1. ALC Men u 6-3

6-2. F requency Men u 6-5

6-3. Marker Men u 6-7

6-4. Mo dulation Men u 6-9

6-5. P ow er Men u 6-11

6-6. Service Men u 6-13

6-7. Sw eep Men u 6-15

6-8. System Men u 6-17

6-9. User Cal Men u 6-19

7-1. Cable Routing Lo cations 7-3

7-2. F ront P anel Disassem bly Flow c hart 7-5

7-3. F ront Edge T rim Strip 7-6

7-4. F ront P anel Screw Lo cations 7-7

7-5. F ront P anel Coaxial and Ribb on Cable Lo cations 7-8

7-6. F ront P anel Pro cessor Cables 7-9

7-7. Displa y Remo v al 7-10

7-8. Source Mo dule In terface Remo v al 7-11

7-9. RPG Remo v al 7-12

7-10. Remo ving RF Output Connector (F ront P anel) 7-13

Contents-13 · Service

Page 27 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 27 · click to enlarge
Page 27

7-11. F ront P anel BNC Lo cations 7-14

7-12. Keyb oard Assem bly Remo v al 7-15

7-13. Keypad and An ti-ro c k Sheets 7-16

7-14. Display Filter 7-17

7-15. Keyb oard Reassem bly 7-18

7-16. Keyb oard Assem bly Screw Lo cations 7-19

7-17. Rear P anel Disassembly Flo w c hart 7-22

7-18. B1 Rear P anel F an Assem bly 7-24

7-19. Rear P anel BNC Connector Lo cations 7-25

7-20. Remo ving RF Output Connector 7-26

7-21. Rear P anel Remo v al 7-27

7-22. Rear P anel Cable Lo cations 7-28

7-23. Reference Standard Moun tings 7-29

7-24. T ransformer Remo v al 7-30

7-25. A19 Rear P anel In terface Remo v al 7-31

7-26. V oltage Selection Switch Remo v al 7-32

7-27. F use, Line Filter, and Ground Wires 7-33

7-28. Rear P anel F use Housing 7-34

7-29. Line Filter Remo v al 7-35

7-30. Option 004 BNC Connectors Remo v al 7-36

7-31. RF Dec k Disassembly Flow c hart 7-38

7-32. Remo ving RF Output Connector 7-40

7-33. W51 RF Cable Lo cation 7-41

7-34. RF Dec k in the Service P osition 7-42

7-35. A20 RF In terface 7-43

7-36. RF Dec k Cable Lo cations { 83640B /L and 83650B /L 7-45

7-37. RF Dec k Cable Lo cations { 83640B /L Option 001 and 83650B /L Option 001 7-46

7-38. RF Dec k Cable Lo cations { 83640B /50B Option 006 7-47

7-39. RF Dec k Cable Lo cations { 83640B /50B Options 001 and 006 7-48

7-40. RF Dec k Cable Lo cations { 83620B /22B and 83630B /L 7-49

7-41. RF Dec k Cable Lo cations { 83620B /22B and 83630B /L Option 001 7-50

7-42. RF Dec k Cable Lo cations { 83620B /22B/30B Option 006 7-51

7-43. RF Dec k Cable Lo cations { 83620B /22B/30B Options 001 and 006 7-52

7-44. RF Dec k Cable Lo cations { 83623B 7-53

7-45. RF Dec k Cable Lo cations { 83623B Option 001 7-54

7-46. RF Dec k Cable Lo cations { 83624B 7-57

7-47. RF Dec k Cable Lo cations { 83624B Option 001 7-58

7-48. RF Dec k Cable Lo cations { 83623B /24B Option 006 7-59

7-49. RF Dec k Cable Lo cations { 83623B /24B Options 001 and 006 7-60

7-50. RF Dec k Screw Lo cations 7-61

7-51. Motherb oard Disassembly Flow c hart 7-63

7-52. F ront Edge T rim Strip 7-64

7-53. F ront P anel Screw Lo cations 7-65

7-54. F ront P anel Coaxial and Ribb on Cable Lo cations 7-66

7-55. Remo ving RF Output Connector (F ront P anel) 7-67

7-56. Remo ving RF Output Connector (Option 004) 7-69

7-57. Rear P anel Remo v al 7-70

7-58. Rear P anel Cable Lo cations 7-71

7-59. Ma jor Assem blies 7-72

7-60. W51 Lo cation 7-73

7-61. Motherb oard Cables 7-74

Contents-14 Agilent 8360 · Service

Page 28 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 28 · click to enlarge
Page 28

7-62. Motherb oard Screws 7-75

9-1. Mo dule Exc hange Program 9-4

9-2. Ma jor Assem blies (1 of 5) 9-5

9-2. Ma jor Assem blies (2 of 5) 9-6

9-2. Ma jor Assem blies (3 of 5) 9-7

9-2. Ma jor Assem blies (4 of 5) 9-8

9-2. Ma jor Assem blies (5 of 5) 9-9

9-3. Cables (1 of 19) 9-15

9-3. Cables (2 of 19) 9-16

9-3. Cables (3 of 19) 9-17

9-3. Cables (4 of 19) 9-18

9-3. Cables (5 of 19) 9-19

9-3. Cables (6 of 19) 9-20

9-3. Cables (7 of 19) 9-21

9-3. Cables (8 of 19) 9-22

9-3. Cables (9 of 19) 9-23

9-3. Cables (10 of 19) 9-24

9-3. Cables (11 of 19) 9-25

9-3. Cables (12 of 19) 9-26

9-3. Cables (13 of 19) 9-27

9-3. Cables (14 of 19) 9-28

9-3. Cables (15 of 19) 9-29

9-3. Cables (16 of 19) 9-30

9-3. Cables (17 of 19) 9-31

9-3. Cables (18 of 19) 9-32

9-3. Cables (19 of 19) 9-33

9-4. F ront P anel Hardware 9-40

9-5. F ront P anel, Casting and Keypad 9-42

9-6. F ront P anel, A ttaching Hardware (1 of 3) 9-43

9-6. F ront P anel, A ttaching Hardware (2 of 3) 9-44

9-6. F ront P anel, A ttaching Hardware (3 of 3) 9-45

9-7. Rear P anel Hardware 9-46

9-8. Rear P anel F an Assem bly 9-49

9-9. Bottom View, A ttaching Hardware 9-50

9-10. T op View, A ttaching Hardware 9-51

9-11. Left Side View, A ttaching Hardware 9-53

9-12. Righ t Side View, A ttaching Hardware 9-54

9-13. RF Dec k Righ t Side View, A ttaching Hardware 9-55

9-14. RF Connector and A ttaching Hardware 9-56

9-15. Chassis P arts 9-57

10-1. Remo ving the F an Filter 10-3

10-2. Replacing the Line F use 10-5

10-3. A15BT1 Battery Lo cation 10-6

12-3. Ma jor Assem bly and Cable Lo cations 12-3

12-4. A4L1 Lo cation 12-4

12-5. F ractional-N Reference and API Spur Adjustmen t Setup 12-6

12-6. F ractional-N Reference and API Spurs Adjustmen t Lo cations 12-7

Contents-15 · Service

Page 29 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 29 · click to enlarge
Page 29

2-1. Op eration V eri cation F orm 2-4

2-2. Sw ept F requency Accuracy Instrumen t Settings 2-14

2-3. Additional Instrumen t Settings 2-15

2-4. Additional Instrumen t Settings 2-16

2-5. Additional Instrumen t Settings 2-16

2-6. Additional Instrumen t Settings 2-17

2-7. CW F requency Switching Time Settings 2-19

2-8. P ow er Accuracy F requencies 2-24

2-9. Start and Stop F requencies 2-33

2-10. Corresp onding YO F requency Ranges and RF Output F requencies 2-33

2-11. Corresp onding YO F requency Ranges and RF Output F requencies 2-36

2-12. Determining the Actual Spur F requencies 2-42

2-13. Lo w Band Spurious Signals 2-43

2-14. Sp ectrum Analyzer F requencies and Line Spur F requencies and Harmonics . 2-46

2-15. Phase Noise Measuremen t System P arameters 2-50

2-16. F requency Setting for Phase Noise Measuremen ts 2-51

2-17. Pulse On/O Cen ter F requencies 2-53

2-18. Syn thesizer F requencies 2-56

2-19. Syn thesizer and LO F requencies 2-62

2-20. Risetime and F alltime V alues 2-68

2-21. Lo w Band Video F eedthrough F requencies 2-74

2-22. P ow er Lev el Con v ersions to V olts P eak 2-76

2-23. Millimeter Band Video F eedthrough F requencies 2-77

2-24. Syn thesizer F requencies 2-85

2-25. P ow er Meter Readings 2-89

2-26. RF Signal Lev el 2-90

2-27. Syn thesizer F requencies 2-93

2-28. Syn thesizer F requencies 2-98

2-29. DC FM T est P ass-F ail Results 2-101

2-30. FM Deviation F requencies and Settings 2-103

2-33. T est Record for Agilent 83620B and 83622B (1 of 8) 2-109

2-33. T est Record for 83620B and 83622B (2 of 8) 2-110

2-33. T est Record for 83620B and 83622B (3 of 8) 2-111

2-33. T est Record for 83620B and 83622B (4 of 8) 2-112

2-33. T est Record for 83620B and 83622B (5 of 8) 2-113

2-33. T est Record for 83620B and 83622B (6 of 8) 2-114

2-33. T est Record for 83620B and 83622B (7 of 8) 2-115

2-33. T est Record for 83620B and 83622B (8 of 8) 2-116

2-34. T est Record for Agilent 83623B and 83624B (1 of 8) 2-117

2-34. T est Record for 83623B and 83624B (2 of 8) 2-118

2-34. T est Record for 83623B and 83624B (3 of 8) 2-119

2-34. T est Record for 83623B and 83624B (4 of 8) 2-120

2-34. T est Record for 83623B and 83624B (5 of 8) 2-121

T ables · Contents-16 Agilent 8360 · Service

Page 30 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 30 · click to enlarge
Page 30

2-34. T est Record for 83623B and 83624B (6 of 8) 2-122

2-34. T est Record for 83623B and 83624B (7 of 8) 2-123

2-34. T est Record for 83623B and 83624B (8 of 8) 2-124

2-35. T est Record for Agilent 83630B (1 of 8) 2-125

2-35. T est Record for 83630B (2 of 8) 2-126

2-35. T est Record for 83630B (3 of 8) 2-127

2-35. T est Record for 83630B (4 of 8) 2-128

2-35. T est Record for 83630B (5 of 8) 2-129

2-35. T est Record for 83630B (6 of 8) 2-130

2-35. T est Record for 83630B (7 of 8) 2-131

2-35. T est Record for 83630B (8 of 8) 2-132

2-36. T est Record for Agilent 83640B (1 of 8) 2-133

2-36. T est Record for 83640B (2 of 8) 2-134

2-36. T est Record for 83640B (3 of 8) 2-135

2-36. T est Record for the 83640B (4 of 8) 2-136

2-36. T est Record for the 83640B (5 of 8) 2-137

2-36. T est Record for the 83640B (6 of 8) 2-138

2-36. T est Record for the 83640B (7 of 8) 2-139

2-36. T est Record for the 83640B (8 of 8) 2-140

2-37. T est Record for Agilent 83650B (1 of 8) 2-141

2-37. T est Record for 83650B (2 of 8) 2-142

2-37. T est Record for 83650B (3 of 8) 2-143

2-37. T est Record for 83650B (4 of 8) 2-144

2-37. T est Record for 83650B (5 of 8) 2-145

2-37. T est Record for 83650B (6 of 8) 2-146

2-37. T est Record for 83650B (7 of 8) 2-147

2-37. T est Record for 83650B (8 of 8) 2-148

2-34. T est Record for Agilent 83623L (1 of 5) 2-149

2-34. T est Record for 83623L (2 of 5) 2-150

2-34. T est Record for 83623L (3 of 5) 2-151

2-34. T est Record for 83623L (4 of 5) 2-152

2-34. T est Record for 83623L (5 of 5) 2-153

2-35. T est Record for Agilent 83630L (1 of 5) 2-154

2-35. T est Record for 83630L (2 of 5) 2-155

2-35. T est Record for 83630L (3 of 5) 2-156

2-35. T est Record for 83630L (4 of 5) 2-157

2-35. T est Record for 83630L (5 of 5) 2-158

2-36. T est Record for Agilent 83640L (1 of 5) 2-159

2-36. T est Record for 83640L (2 of 5) 2-160

2-36. T est Record for 83640L (3 of 5) 2-161

2-36. T est Record for the 83640L (4 of 5) 2-162

2-36. T est Record for the 83640L (5 of 5) 2-163

2-37. T est Record for Agilent 83650L (1 of 6) 2-164

2-37. T est Record for 83650L (2 of 6) 2-165

2-37. T est Record for 83650L (3 of 6) 2-166

2-37. T est Record for 83650L (4 of 6) 2-167

2-37. T est Record for 83650L (5 of 6) 2-168

2-37. T est Record for 83650L (6 of 6) 2-169

3-1. Calibration Constant Descriptions 3-22

3-2. Start and Stop F requencies (GHz) 3-31

3-4. Ampli er/Filter Calibration Constants and Default V alues 3-39

Contents-17 · Service

Page 31 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 31 · click to enlarge
Page 31

3-5. F requency Ranges 3-50

3-6. ALC O set Calibration Constant Default V alues 3-63

3-7. Cen ter F requencies and Calibration Constants 3-65

3-8. ALC Calibration Constants and Default V alues 3-68

3-9. P ow er Lev el and Calibration Constant Adjustmen t 3-69

3-10. P ow er Lev el and Calibration Constant Adjustmen t 3-70

4-9. Calibration Constant Descriptions 4-9

4-10. Hardware Con guration Cal Constants #463 and #464 4-32

4-11. Assem blies and Corresp onding Cal Constant ID Num b ers and Revisions . . 4-33

5-1. Required Binaries 5-2

5-2. T est Equipmen t Required 5-5

5-3. T est Equipmen t Required 5-10

5-4. Syn thesizer Hardk ey Em ulation 5-18

8-1. Adjustmen ts and P erformance T ests Required After Repair or Replacemen t of

an Assem bly 8-1

9-1. Reference Designations and Abbreviations 9-3

9-2. Ma jor Assem blies 9-10

9-3. Cables 9-34

9-4. F ront P anel Hardware 9-41

9-5. F ront P anel, Casting and Keypad 9-42

9-6. F ront P anel, A ttaching Hardware 9-43

9-7. Rear P anel (1 of 2) 9-47

9-7. Rear P anel (2 of 2) 9-48

9-8. Rear P anel F an Assem bly 9-49

9-9. Bottom View, A ttac hing Hardware 9-50

9-10. T op View, A ttaching Hardware 9-52

9-11. Left Side View, A ttaching Hardware 9-53

9-12. Righ t Side View, A ttac hing Hardware 9-54

9-13. RF Dec k Righ t Side View, A ttaching Hardware 9-55

9-14. RF Connector and A ttaching Hardware 9-56

9-15. Chassis P arts 9-58

9-16. Miscellaneous Replaceable Accessories 9-59

12-1 12-2

12-2. Summary of F ractional-N Reference & API Spurs Adjustmen t 12-5

12-3. Assem blies and Corresp onding Cal Constant ID Num b ers and Revisions . . 12-9

12-4. Adjustmen ts and P erformance T ests Required After Repair or Replacemen t of

an Assem bly 12-9

Contents-18 Agilent 8360 · Service

Page 32 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 32 · click to enlarge
Page 32

Eac h R e quir e d Equipment list is sho wn b elow. The page n um b er of eac h list is also included

for easy access.

Agilent 83620B /22B Required Equipmen t : 1-2

Agilent 83623B /24B Required Equipmen t : 1-9

Agilent 83630B Required Equipmen t : 1-16

Agilent 83640B Required Equipmen t : 1-23

Agilent 83650B Required Equipmen t : 1-30

Agilent 83623L Required Equipmen t : 1-38

Agilent 83630L Required Equipmen t : 1-43

Agilent 83640L Required Equipmen t : 1-48

Agilent 83650L Required Equipmen t : 1-53

Agilent 8360 Required Equipment

Equipment Required · Agilent 83620B/22B

Page 33 of the Agilent Technologies 83640B user manual — ADC (AA)
Page 33 · click to enlarge
Page 33

Agilent 83620B/22B Required Equipment

T able 1-1 Required Equipment for 83620B/22B

Instrument Critical Sp eci cations Recommended

F requency Range10 MHz to 20 GHz
F requency Span0 Hz, 100 Hz to 20 GHz
Amplitude Range+20 to 100 dBm
Resolution Bandwidth10 Hz to 3 MHz
Video Bandwidth10 Hz to 3 MHz
Log Fidelity 0.1 dB/dB o v er 0 to

80 dB display (1.0 dB maxim um)

Video OutputDC v oltage prop ortional

to v ertical p osition of trace on display

Pulse Mo dulation On/O

Capable of phase lo cking to external

AM Dynamic Range (P)

Maxim um FM Deviation (P)

and API Spurs (A)

Square W a v e Symmetry (A)

A vs B Sweep Mo de

V ertical Sensitivit y5 mV/Div
Bandwidth100 MHz

Pulse P erformance-Alt (P)

F requency Range20 Hz to 10 MHz HP/Agilent 3585A/B AM Bandwidth (P)

Controller 4 Mb yte RAM

Step A ttenuator Flatness (AP)

BASIC 5.1

P o wer Flatness and

GPIB

Step A ttenuator Flatness (AA)

ADC (AA)

P o wer Flatness (AA)

1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Show 44 more labels / short text on this page
Use · Mo del · Sp ectrum · HP/Agilent 8566B · Swept F requency · Analyzer · Accuracy (P) · Spurious Signals · (Harmonic) (P) · Spurious Signals · (Non-harmonic) (P) · Spurious Signals · (Line Related) (P) · Ratio (P) · 10 MHz reference · FM Accuracy (P) · F ractional{N Reference · FM Gain (A) · Analog · HP/Agilent 1740A/ · Amp/Mult Adjustments (A) · Oscilloscop e · HP/Agilent 54600B · Amp/Filt Adjustments (A) · Lo cal · 83620/ · Oscillator · HP/Agilent 8340A/B · AM Accuracy (P) · (Synthesized · Sweep er) · Sp ectrum · Analyzer with · FM Bandwidth (P) · T rac king · Generator · HP/Agilent 9836/ · HP/Agilent 9920/ · HP/Agilent 310/ · Accuracy (AP) · HP/Agilent 320 · YO Delay (AA) · 1-2 Required Equipment · Agilent 83620B/22B
Page 34 of the Agilent Technologies 83640B user manual — ADC (AA)
Page 34 · click to enlarge
Page 34

T able 1-1. Required Equipment for 83620B/22B (continued)

Instrument Critical Sp eci cations Recommended

P/N 08360-10001 Step A ttenuator Flatness (AP)

(Shipp ed with instrument)

P o wer Flatness and

Step A ttenuator Flatness (AA)

ADC (AA)

P o wer Flatness (AA)

D VM Range 50 to +50 VDC
Accuracy 0.01%
Input Imp edance 10 M

AM Dynamic Range (P)

Mo dulation Meter (P)

Lo w P o wer SRD Bias (A)

ADC (AA)

In ternal TimebaseAging
BandwidthDC to 300 MHz
Input Imp edance1 M and 50
V ertical Sensitivit y 5 mV/Div
Horizon tal Sensitivit y50 ns/Div

F requency Switc hing Time (P)

T riggerEv ent T riggerable

Pulse P erformance-Alt (P)

Pulse Mo dulation Video

In ternal Pulse Accuracy (P)

10 MHz Standard (A)

Mo dulator O set and Gain (A)

3 dB Bandwidth20 GHz

Pulse P erformance (P)

DC V oltage Accuracy 0.4% of full

scale 2 mV

T rigger Sensitivit y DC to 100 MHz:

T rigger & Timebase Jitter:

2.5 ps + 5E

Time In terv al Accuracy:

10 ps 0.1% of reading

Channel Input Re ection:

5% for 30 ps risetime

1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Agilent 8360 Required Equipment

Show 39 more labels / short text on this page
Use · Mo del · Software No Substitute · Accuracy (AP) · YO Delay (AA) · HP/Agilen t 3456A · / · External Leveling (P) · HP/Agilen t 3457A · AM Accuracy (P) · AM Bandwidth (P) · FM Bandwidth (P) · AM Accuracy (A) · AM Delay (A) · Digitizing · Dual Channel · HP/Agilen t 54110A/ · Oscilloscop e · HP/Agilen t 54111D · / · Rate (P) · HP/Agilen t 54222A/ · Swept F requency · HP/Agilen t 54750A/ · Accuracy (P) · HP/Agilen t 54752A · F eedthrough (P) · FM Bandwidth (P) · Mo dulation Generator · Flatness (A) · Digitizing · HP/Agilen t 54124T/ · Oscilloscop e · HP/Agilen t 54750A/ · HP/Agilen t 54752A · 40 mV p-p · 5 · x delay setting · Agilent 83620B/22B
Page 35 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 35 · click to enlarge
Page 35

T able 1-1. Required Equipment for 83620B/22B (continued)

Instrument Critical Sp eci cations Recommended

Division Ratio1:1 HP/Agilent 10437A In ternal Pulse Accuracy (P)
Division Ratio10:1 HP/Agilent 10431A Swept F requency

F requency Switching Time (P)

Pulse Mo dulation Video

Mo dulator O set and Gain (A)

Pulse Width 50 ns

Pulse P erformance (P)

Rise Time 10 ns

Pulse Mo dulation Video

F requency10 Hz to 5 MHz
F requency Accuracy 5 x 10

HP/Agilent 3325A/B AM Accuracy (P)

AM Dynamic Range (P)

100 kHz to 1 MHz 0.1 dB
100 kHz to 20 MHz 0.4 dB

Maxim um FM Deviation (P)

Mo dulation Meter (P)

Amplitude Range> 16 V p-p HP/Agilent 8111A/

Maxim um FM Deviation (P)

P o wer Meter P o wer Range: 1 W to 100 mW

P o wer Accuracy (P)

Accuracy 0.02 dB

AM Dynamic Range (P)

ALC P o wer Level

P o wer Meter P o wer Range: 1 W to 100 mW

HP/Agilent 437B P o wer Flatness (P)

Accuracy 0.02 dB

P o wer Flatness (A)

1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Show 35 more labels / short text on this page
Use · Mo del · Oscilloscop e · Prob es · Oscilloscop e · Prob es · Accuracy (P) · F eedthrough (P) · Amp/Mult Adjustments (A) · Amp/Filt Adjustments (A) · Pulse · HP/Agilent 8112B/ · Generator · HP/Agilent 8116A · F eedthrough (P) · 6 · F unction · Generator · Amplitude Accuracy · FM Accuracy (P) · AM Accuracy (A) · AM Delay (A) · FM Gain (A) · F unction · Generator · HP/Agilent 8116A/ · HP/Agilent 8175A · HP/Agilent 436A/ · HP/Agilent 437B/ · HP/Agilent 438A · Amp/Filt Adjustments (A) · Accuracy (A) · AM Accuracy (A) · 1-4 Required Equipment · Agilent 83620B/22B
Page 36 of the Agilent Technologies 83640B user manual — AM
Page 36 · click to enlarge
Page 36

T able 1-1. Required Equipment for 83620B/22B (continued)

Instrument Critical Sp eci cations Recommended

P o wer Meter P o wer Range: 1 W to 100 mW

HP/Agilen t 438A P o wer Flatness and

Accuracy 0.02 dB

P o wer Flatness (AA)

Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AP)

P o wer Sensor F requency Range: 10 MHz to 2.3 GHz

HP/Agilen t 8482A P o wer Flatness (P)

P o wer Range1 W to 100 mW

P o wer Flatness (A)

Step A ttenuator Flatness (AP)

P o wer Flatness and

Step A ttenuator Flatness (AA)

P o wer Flatness (AA)

P o wer Sensor F requency Range: 50 MHz to 20 GHz

HP/Agilen t 8485D Step A ttenuator Flatness (AA)

P o wer Range100 pW to 10 W

Step A ttenuator Flatness (AP)

P o wer Sensor F requency Range: 50 MHz to 20 GHz

HP/Agilen t 8485A P o wer Accuracy (P)

P o wer Range1 W to 100 mW

P o wer Flatness (P)

AM Dynamic Range (P)

ALC P o wer Level

P o wer Flatness (A)

P o wer Flatness and

P o wer Flatness (AA)

Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AP)

A ttenuator A ttenuation: 30 dB 0.05 dB at 50 MHz HP/Agilen t 11708A Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AP)

F requency Range (tuned):

2.5 MHz to 1.3 GHz

Range0 dBm to 127 dBm
Relative P o wer Accuracy 0.5 dB

AM

Rates20 Hz to 100 kHz
Depthto 99%
Accuracy 1% of reading 1 count

1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Agilent 8360 Required Equipment

Show 13 more labels / short text on this page
Use · Mo del · Accuracy (AP) · Accuracy (AP) · Amp/Filt Adjustments (A) · Accuracy (A) · AM Accuracy (A) · Accuracy (AP) · Measuring · HP/Agilen t 8902A · AM Accuracy (P) · Receiv er · Agilent 83620B/22B
Page 37 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 37 · click to enlarge
Page 37

T able 1-1. Required Equipment for 83620B/22B (continued)

Instrument Critical Sp eci cations Recommended

F requency Range (carrier):

HP/Agilent 3048A Single Sideband Phase

0.01 to 18 GHz

Sensitivit y:

< 70 dBc at 100 Hz o set

< 78 dBc at 1 kHz o set

< 86 dBc at 10 kHz o set

< 107 dBc at 100 kHz o set

O set F requency Range:

100 Hz to 2 MHz

Amplitude Accuracy:

2 dB to 1 MHz o set

F requency10 MHz
In ternal TimebaseAging
Stability< 1 x 10

10 MHz Standard (A)

F requency Range1.5 to 18 GHz

HP/Agilent 8348A FM Bandwidth (P)

Leveled Output P o wer 16 dBm
F requency Range100 kHz to 1.3 GHz

HP/Agilent 8447F Pulse P erformance-Alt (P)

Preampli er Gain25 dB

Pulse Mo dulation Video

P o wer Ampli er Gain: 22 dB

1 meter of coax or

Mixer F requency Range1 GHz to 20 GHz

P/N 0955-0307 Pulse P erformance-Alt (P)

P o wer Splitter F requency Range: 10 MHz to 20 GHz HP/Agilent 11667B FM Bandwidth (P)

F requency RangeDC to 20 GHz

HP/Agilent 33330D External Leveling (P)

Maxim um Input200 mW
P olarit yNegativ e
A ttenuator F requency Range10 MHz to 20 GHz

Pulse P erformance-Alt (P)

Maxim um Input P o wer: 300 mW

Pulse Mo dulation Video

A ttenuation6 dB

1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est

P { Manual P erformance T est

Show 34 more labels / short text on this page
Use · Mo del · Phase Noise · Measuremen t · Noise (P) · System · F requency · HP/Agilent 5061A/ · Standard · 10 · /yr · HP/Agilent 5071A · Rate (P) · Microwa v e · Ampli er · Preampli er/ · P o wer · Ampli er · F eedthrough (P) · Delay Line · FM Bandwidth (P) · Discriminator · semirigid cable · (Norda 4817) · AM Accuracy (P) · FM Bandwidth (P) · Crystal · Detector · AM Bandwidth (P) · HP/Agilent 8493C · Opt 006 · F eedthrough (P) · 1-6 Required Equipment · Agilent 83620B/22B
Page 38 of the Agilent Technologies 83640B user manual — SLP-550
Page 38 · click to enlarge
Page 38

T able 1-1. Required Equipment for 83620B/22B (continued)

Instrument Critical Sp eci cations Recommended

A ttenuator F requency Range10 MHz to 20 GHz

P o wer Accuracy (P)

Maxim um Input P o wer: 300 mW

A ttenuation10 dB

Pulse P erformance (P)

Pulse P erformance-Alt (P)

FM Bandwidth (P) (2)

Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AA)

A ttenuator F requency Range10 MHz to 20 GHz

Maxim um Leveled P o wer (P)

Maxim um Input P o wer: 300 mW

Pulse P erformance (P)

A ttenuation20 dB

3.7 GHz Lo w

P/N 9135-0191 Pulse Mo dulation Video

K & L Microwa v e

Pulse Mo dulation Video

5LL30-130/BT2400/BP

500 MHz Lo w

Pulse Mo dulation Video

SLP-550

T o ol Kit No Substitute P/N 08360{60060

In v ertron California

Capacitor 1000 pf P/N 0160-4574 External Leveling (P)

1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est

P { Manual P erformance T est

Agilent 8360 Required Equipment

Show 24 more labels / short text on this page
Use · Mo del · HP/Agilen t 8493C · Opt 010 · Spurious Signals · (Harmonic) (P) · AM Accuracy (P) · Amp/Mult Adjustments (A) · Amp/Filt Adjustments (A) · HP/Agilen t 8493C · Option 020 · P ass Filter · F eedthrough (P) · 130 MHz · Bessel Lo w · F eedthrough (P) · P ass Filter · Mini-Circuits · P ass Filter · F eedthrough (P) · Spurious Signals · Instruments 501TC · (Line Related) (P) · Agilent 83620B/22B
Page 39 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 39 · click to enlarge
Page 39

The following list of adapters and cables is provided for con v enience. They may b e used in

equipmen t setups for p erformance tests or adjustmen ts.

SMA (m) to SMA (m) adapter 1250-1159

SMA (f ) to SMA (f ) adapter 1250-1158

SMB (m) to SMB (m) adapter 1250-0669

SMB (f ) to SMB (f ) adapter 1250-0672

SMB tee (f ) (m) (m) 1250-1391

3.5 mm (f ) to 3.5 mm (f ) adapter 5061-5311

3.5 mm (f ) to N-t yp e (m) adapter 1250-1745

2.4 mm (f ) to 2.92 mm (f ) adapter 1250-2187

2.4 mm (f ) to 2.4 mm (f ) adapter 1250-2188

2.4 mm (m) to 3.5 mm (f ) adapter 11901C

BNC (f ) to BNC (f ) adapter 1250-0080

BNC (m) to BNC (m) adapter 1250-0216

BNC (f ) to SMA (m) adapter 1250-1200

BNC (f ) to SMB (m) adapter 1250-1237

BNC tee (m) (f ) (f ) 1250-0781

SMB (f ) to BNC (m) exible cable 85680-60093

SMA semi-rigid cable 2 feet 08340-20124

GPIB cable 2 meters HP/Agilent 10833B

BNC cable 8120-2582 · 1-8 Required Equipment · Agilent 83620B/22B

Page 40 of the Agilent Technologies 83640B user manual — ADC (AA)
Page 40 · click to enlarge
Page 40

Agilent 83623B/24B Required Equipment

T able 1-2. Required Equipment for 83623B/24B

Instrument Critical Sp eci cations Recommended

F requency Range10 MHz to 20 GHz
F requency Span0 Hz, 100 Hz to 20 GHz
Amplitude Range+20 to 100 dBm
Resolution Bandwidth10 Hz to 3 MHz
Video Bandwidth10 Hz to 3 MHz
Log Fidelit y 0.1 dB/dB o v er 0 to

80 dB display (1.0 dB maxim um)

Video OutputDC v oltage prop ortional

to v ertical p osition of trace on display

Pulse Mo dulation On/O

Capable of phase lo cking to external

AM Dynamic Range (P)

Maxim um FM Deviation (P)

and API Spurs (A)

Square W a v e Symmetry (A)

A vs B Sweep Mo de

V ertical Sensitivit y5 mV/Div
Bandwidth100 MHz

Pulse P erformance-Alt (P)

F requency Range20 Hz to 10 MHz HP/Agilen t 3585A/B AM Bandwidth (P)

Controller 4 Mb yte RAM

Step A ttenuator Flatness (AP)

BASIC 5.1

P o wer Flatness and

GPIB

Step A ttenuator Flatness (AA)

ADC (AA)

P o wer Flatness (AA)

1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Agilent 8360 Required Equipment

Show 43 more labels / short text on this page
Use · Mo del · Sp ectrum · HP/Agilen t 8566B · Swept F requency · Analyzer · Accuracy (P) · Spurious Signals · (Harmonic) (P) · Spurious Signals · (Non-harmonic) (P) · Spurious Signals · (Line Related) (P) · Ratio (P) · 10 MHz reference · FM Accuracy (P) · F ractional{N Reference · FM Gain (A) · Analog · HP/Agilen t 1740A/ · Amp/Mult Adjustments (A) · Oscilloscop e · HP/Agilen t 54600B · Amp/Filt Adjustments (A) · Lo cal · 83620/ · Oscillator · HP/Agilen t 8340A/B · AM Accuracy (P) · (Synthesized · Sweep er) · Sp ectrum · Analyzer with · FM Bandwidth (P) · T rac king · Generator · HP/Agilen t 9836/ · HP/Agilen t 9920/ · HP/Agilen t 310/ · Accuracy (AP) · HP/Agilen t 320 · YO Delay (AA) · Agilent 83623B/24B
Page 41 of the Agilent Technologies 83640B user manual — ADC (AA)
Page 41 · click to enlarge
Page 41

T able 1-2. Required Equipment for 83623B/24B (continued)

Instrument Critical Sp eci cations Recommended

P/N 08360-10001 Step A ttenuator Flatness (AP)

(Shipp ed with instrument)

P o wer Flatness and

Step A ttenuator Flatness (AA)

ADC (AA)

P o wer Flatness (AA)

D VM Range 50 to +50 VDC
Accuracy 0.01%
Input Imp edance 10 M

AM Dynamic Range (P)

Mo dulation Meter (P)

Lo w P o wer SRD Bias (A)

ADC (AA)

In ternal TimebaseAging
BandwidthDC to 300 MHz
Input Imp edance1 M and 50
V ertical Sensitivit y 5 mV/Div
Horizon tal Sensitivit y50 ns/Div

F requency Switching Time (P)

T riggerEv ent T riggerable

Pulse P erformance-Alt (P)

Pulse Mo dulation Video

In ternal Pulse Accuracy (P)

10 MHz Standard (A)

Mo dulator O set and Gain (A)

3 dB Bandwidth20 GHz

Pulse P erformance (P)

DC V oltage Accuracy 0.4% of full

scale 2 mV

T rigger Sensitivit y DC to 100 MHz:

T rigger & Timebase Jitter:

2.5 ps + 5E

Time In terv al Accuracy:

10 ps 0.1% of reading

Channel Input Re ection:

5% for 30 ps risetime

Division Ratio1:1 HP/Agilent 10437A In ternal Pulse Accuracy (P)

1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Show 42 more labels / short text on this page
Use · Mo del · Software No Substitute · Accuracy (AP) · YO Delay (AA) · HP/Agilent 3456A · / · External Leveling (P) · HP/Agilent 3457A · AM Accuracy (P) · AM Bandwidth (P) · FM Bandwidth (P) · AM Accuracy (A) · AM Delay (A) · Digitizing · Dual Channel · HP/Agilent 54110A/ · Oscilloscop e · Rate (P) · HP/Agilent 54111D · / · HP/Agilent 54222A/ · Swept F requency · HP/Agilent 54750A/ · Accuracy (P) · HP/Agilent 54752A · F eedthrough (P) · FM Bandwidth (P) · Mo dulation Generator · Flatness (A) · Digitizing · HP/Agilent 54124T/ · Oscilloscop e · HP/Agilent 54750A/ · HP/Agilent 54752A · 40 mV p-p · 5 · x delay setting · Oscilloscop e · Prob es · 1-10 Required Equipment · Agilent 83623B/24B
Page 42 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 42 · click to enlarge
Page 42

T able 1-2. Required Equipment for 83623B/24B (continued)

Instrument Critical Sp eci cations Recommended

Division Ratio10:1 HP/Agilen t 10431A Swept F requency

F requency Switc hing Time (P)

Pulse Mo dulation Video

Mo dulator O set and Gain (A)

Pulse Width 50 ns

Pulse P erformance (P)

Rise Time 10 ns

Pulse Mo dulation Video

F requency10 Hz to 5 MHz
F requency Accuracy 5 x 10

HP/Agilen t 3325A/B AM Accuracy (P)

AM Dynamic Range (P)

100 kHz to 1 MHz 0.1 dB
100 kHz to 20 MHz 0.4 dB

Maxim um FM Deviation (P)

Mo dulation Meter (P)

Amplitude Range> 16 V p-p HP/Agilen t 8111A/

Maxim um FM Deviation (P)

P o wer Meter P o wer Range: 1 W to 100 mW

P o wer Accuracy (P)

Accuracy 0.02 dB

AM Dynamic Range (P)

ALC P o wer Level

P o wer Meter P o wer Range: 1 W to 100 mW

HP/Agilen t 437B P o wer Flatness (P)

Accuracy 0.02 dB

P o wer Flatness (A)

P o wer Meter P o wer Range: 1 W to 100 mW

HP/Agilen t 438A P o wer Flatness and

Accuracy 0.02 dB

P o wer Flatness (AA)

Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AP)

1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Agilent 8360 Required Equipment

Show 33 more labels / short text on this page
Use · Mo del · Oscilloscop e · Prob es · Accuracy (P) · F eedthrough (P) · Amp/Mult Adjustments (A) · Amp/Filt Adjustments (A) · Pulse · HP/Agilen t 8112B/ · Generator · HP/Agilen t 8116A · F eedthrough (P) · F unction · 6 · Generator · Amplitude Accuracy · FM Accuracy (P) · AM Accuracy (A) · AM Delay (A) · FM Gain (A) · F unction · Generator · HP/Agilen t 8116A/ · HP/Agilen t 8175A · HP/Agilen t 436A/ · HP/Agilen t 437B/ · HP/Agilen t 438A · Amp/Filt Adjustments (A) · Accuracy (A) · AM Accuracy (A) · Accuracy (AP) · Agilent 83623B/24B
Page 43 of the Agilent Technologies 83640B user manual — AM
Page 43 · click to enlarge
Page 43

T able 1-2. Required Equipment for 83623B/24B (continued)

Instrument Critical Sp eci cations Recommended

P o wer Sensor F requency Range: 10 MHz to 2.3 GHz

HP/Agilent 8482A P o wer Flatness (P)

P o wer Range1 W to 100 mW

P o wer Flatness (A)

Step A ttenuator Flatness (AP)

P o wer Flatness and

Step A ttenuator Flatness (AA)

P o wer Flatness (AA)

P o wer Sensor F requency Range: 50 MHz to 20 GHz

HP/Agilent 8485D Step A ttenuator Flatness (AA)

P o wer Range100 pW to 10 W

Step A ttenuator Flatness (AP)

P o wer Sensor F requency Range: 50 MHz to 20 GHz

HP/Agilent 8485A P o wer Accuracy (P)

P o wer Range1 W to 100 mW

P o wer Flatness (P)

AM Dynamic Range (P)

ALC P o wer Level

P o wer Flatness (A)

P o wer Flatness and

P o wer Flatness (AA)

Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AP)

A ttenuator A ttenuation: 30 dB 0.05 dB at 50 MHz HP/Agilent 11708A Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AP)

F requency Range (tuned):

HP 8902A

2.5 MHz to 1.3 GHz

Range0 dBm to 127 dBm
Relative P o wer Accuracy 0.5 dB

AM

Rates20 Hz to 100 kHz
Depthto 99%
Accuracy 1% of reading 1 count

F requency Range (carrier):

HP 3048A Single Sideband Phase

0.01 to 18 GHz

Sensitivit y:

< 70 dBc at 100 Hz o set

< 78 dBc at 1 kHz o set

< 86 dBc at 10 kHz o set

< 107 dBc at 100 kHz o set

O set F requency Range:

100 Hz to 2 MHz

Amplitude Accuracy:

2 dB to 1 MHz o set

1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Show 16 more labels / short text on this page
Use · Mo del · Accuracy (AP) · Amp/Filt Adjustments (A) · Accuracy (A) · AM Accuracy (A) · Accuracy (AP) · Measuring · AM Accuracy (P) · Receiver · Phase Noise · Measuremen t · Noise (P) · System · 1-12 Required Equipment · Agilent 83623B/24B
Page 44 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 44 · click to enlarge
Page 44

T able 1-2. Required Equipment for 83623B/24B (continued)

Instrument Critical Sp eci cations Recommended

F requency10 MHz
In ternal TimebaseAging
Stability< 1 x 10

10 MHz Standard (A)

F requency Range1.5 to 18 GHz

HP/Agilen t 8348A FM Bandwidth (P)

Leveled Output P o wer 16 dBm
F requency Range100 kHz to 1.3 GHz

HP/Agilen t 8447F Pulse P erformance-Alt (P)

Preampli er Gain25 dB

Pulse Mo dulation Video

P o wer Ampli er Gain: 22 dB

1 meter of coax or

Mixer F requency Range1 GHz to 20 GHz

P/N 0955-0307 Pulse P erformance-Alt (P)

P o wer Splitter F requency Range: 10 MHz to 20 GHz HP/Agilen t 11667B FM Bandwidth (P)

F requency RangeDC to 20 GHz

HP/Agilen t 33330D External Leveling (P)

Maxim um Input200 mW
P olarit yNegativ e
A ttenuator F requency Range10 MHz to 20 GHz

Pulse P erformance-Alt (P)

Maxim um Input P o wer: 300 mW

Pulse Mo dulation Video

A ttenuation6 dB
A ttenuator F requency Range10 MHz to 20 GHz

P o wer Accuracy (P)

Maxim um Input P o wer: 300 mW

A ttenuation10 dB

Pulse P erformance (P)

Pulse P erformance-Alt (P)

FM Bandwidth (P) (2)

Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AA)

A ttenuator F requency Range10 MHz to 20 GHz

Maxim um Leveled P o wer (P)

Maxim um Input P o wer: 300 mW

Pulse P erformance (P)

A ttenuation20 dB

1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est

P { Manual P erformance T est

Agilent 8360 Required Equipment

Show 38 more labels / short text on this page
Use · Mo del · F requency · HP/Agilen t 5061A/ · 10 · Standard · /yr · HP/Agilen t 5071A · Rate (P) · Microwa v e · Ampli er · Preampli er/ · P o wer · Ampli er · F eedthrough (P) · Delay Line · FM Bandwidth (P) · Discriminator · semi-rigid cable · (Norda 4817) · AM Accuracy (P) · FM Bandwidth (P) · Crystal · Detector · AM Bandwidth (P) · HP/Agilen t 8493C · Opt 006 · F eedthrough (P) · HP/Agilen t 8493C · Opt 010 · Spurious Signals · (Harmonic) (P) · AM Accuracy (P) · Amp/Mult Adjustments (A) · Amp/Filt Adjustments (A) · HP/Agilen t 8493C · Option 020 · Agilent 83623B/24B
Page 45 of the Agilent Technologies 83640B user manual — SLP-550
Page 45 · click to enlarge
Page 45

T able 1-2. Required Equipment for 83623B/24B (continued)

Instrument Critical Sp eci cations Recommended

3.7 GHz Lo w

P/N 9135-0191 Pulse Mo dulation Video

K & L Microwa v e

Pulse Mo dulation Video

5LL30-130/BT2400/BP

500 MHz Lo w

Pulse Mo dulation Video

SLP-550

T o ol Kit No Substitute P/N 08360{60060

In v ertron California

Capacitor 1000 pf P/N 0160-4574 External Leveling (P)

1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est

P { Manual P erformance T est

Show 16 more labels / short text on this page
Use · Mo del · P ass Filter · F eedthrough (P) · 130 MHz · Bessel Lo w · F eedthrough (P) · P ass Filter · Mini-Circuits · P ass Filter · F eedthrough (P) · Spurious Signals · Instruments 501TC · (Line Related) (P) · 1-14 Required Equipment · Agilent 83623B/24B
Page 46 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 46 · click to enlarge
Page 46

The follo wing list of adapters and cables is provided for con v enience. They may b e used in

equipmen t setups for p erformance tests or adjustmen ts.

SMA (m) to SMA (m) adapter 1250-1159

SMA (f ) to SMA (f ) adapter 1250-1158

SMB (m) to SMB (m) adapter 1250-0669

SMB (f ) to SMB (f ) adapter 1250-0672

SMB tee (f ) (m) (m) 1250-1391

3.5 mm (f ) to 3.5 mm (f ) adapter 5061-5311

3.5 mm (f ) to N-t yp e (m) adapter 1250-1745

2.4 mm (f ) to 2.92 mm (f ) adapter 1250-2187

2.4 mm (f ) to 2.4 mm (f ) adapter 1250-2188

2.4 mm (m) to 3.5 mm (f ) adapter 11901C

BNC (f ) to BNC (f ) adapter 1250-0080

BNC (m) to BNC (m) adapter 1250-0216

BNC (f ) to SMA (m) adapter 1250-1200

BNC (f ) to SMB (m) adapter 1250-1237

BNC tee (m) (f ) (f ) 1250-0781

SMB (f ) to BNC (m) exible cable 85680-60093

SMA semi-rigid cable 2 feet 08340-20124

GPB cable 2 meters HP/Agilent 10833B

Agilent 8360 Required Equipment

BNC cable 8120-2582 · Agilent 83623B/24B

Page 47 of the Agilent Technologies 83640B user manual — ADC (AA)
Page 47 · click to enlarge
Page 47

Agilent 83630B Required Equipment

T able 1-3. Required Equipment for 83630B

Instrument Critical Sp eci cations Recommended

F requency Range10 MHz to 22 GHz

(26.5 GHz with external mixers)

F requency Span0 Hz, 100 Hz to 22 GHz
Amplitude Range+20 to 100 dBm
Resolution Bandwidth10 Hz to 3 MHz
Video Bandwidth10 Hz to 3 MHz
Log Fidelity 0.1 dB/dB o v er 0 to

80 dB display (1.0 dB maxim um)

Video OutputDC v oltage prop ortional

Pulse Mo dulation On/O

to v ertical p osition of trace on display

Capable of phase lo cking to external

Pulse P erformance-Alt (P)

AM Dynamic Range (P)

Maxim um FM Deviation (P)

and API Spurs (A)

Square W a v e Symmetry (A)

A vs B Sweep Mo de

V ertical Sensitivit y5 mV/Div
Bandwidth100 MHz

Pulse P erformance-Alt (P)

HP 8340A/B

F requency Range20 Hz to 10 MHz HP/Agilent 3585A/B AM Bandwidth (P)

Controller 4 Mb yte RAM

Step A ttenuator Flatness (AP)

BASIC 5.1

P o wer Flatness and

GPIB

Step A ttenuator Flatness (AA)

ADC (AA)

P o wer Flatness (AA)

1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Show 43 more labels / short text on this page
Use · Mo del · Sp ectrum · HP/Agilent 8566B · Swept F requency · Analyzer · Accuracy (P) · Spurious Signals · (Harmonic) (P) · Spurious Signals · (Non-harmonic) (P) · Spurious Signals · (Line Related) (P) · Ratio (P) · 10 MHz reference · FM Accuracy (P) · F ractional{N Reference · FM Gain (A) · Analog · HP/Agilent 1740A/ · Amp/Mult Adjustments (A) · Oscilloscop e · HP/Agilent 54600B · Amp/Filt Adjustments (A) · Lo cal · 83620/ · Oscillator · AM Accuracy (P) · (Synthesized · Sweep er) · Sp ectrum · Analyzer with · FM Bandwidth (P) · T rac king · Generator · HP/Agilent 9836/ · HP/Agilent 9920/ · HP/Agilent 310/ · Accuracy (AP) · HP/Agilent 320 · YO Delay (AA) · 1-16 Required Equipment · Agilent 83630B
Page 48 of the Agilent Technologies 83640B user manual — ADC (AA)
Page 48 · click to enlarge
Page 48

T able 1-3. Required Equipment for 83630B (continued)

Instrument Critical Sp eci cations Recommended

P/N 08360-10001 Step A ttenuator Flatness (AP)

(Shipp ed with instrument)

P o wer Flatness and

Step A ttenuator Flatness (AA)

ADC (AA)

P o wer Flatness (AA)

D VM Range 50 to +50 VDC
Accuracy 0.01%
Input Imp edance 10 M

AM Dynamic Range (P)

Mo dulation Meter (P)

Lo w P o wer SRD Bias (A)

ADC (AA)

In ternal TimebaseAging
BandwidthDC to 300 MHz
Input Imp edance1 M and 50
V ertical Sensitivit y 5 mV/Div
Horizon tal Sensitivit y50 ns/Div

F requency Switc hing Time (P)

T riggerEv ent T riggerable

Pulse P erformance-Alt (P)

Pulse Mo dulation Video

In ternal Pulse Accuracy (P)

10 MHz Standard (A)

Mo dulator O set and Gain (A)

3 dB Bandwidth26.5 GHz

Pulse P erformance (P)

DC V oltage Accuracy 0.4% of full

scale 2 mV

T rigger Sensitivit y DC to 100 MHz:

T rigger & Timebase Jitter:

2.5 ps + 5E

Time In terv al Accuracy:

10 ps 0.1% of reading

Channel Input Re ection:

5% for 30 ps risetime

Division Ratio1:1 HP/Agilen t 10437A In ternal Pulse Accuracy (P)

1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Agilent 8360 Required Equipment

Show 41 more labels / short text on this page
Use · Mo del · Software No Substitute · Accuracy (AP) · YO Delay (AA) · HP/Agilen t 3456A · / · External Leveling (P) · HP/Agilen t 3457A · AM Accuracy (P) · AM Bandwidth (P) · FM Bandwidth (P) · AM Accuracy (A) · AM Delay (A) · Digitizing · Dual Channel · HP/Agilen t 54110A/ · Oscilloscop e · Rate (P) · HP/Agilen t 54111D · / · HP/Agilen t 54222A/ · Swept F requency · HP/Agilen t 54750A/ · Accuracy (P) · HP/Agilen t 54752A · F eedthrough (P) · FM Bandwidth (P) · Mo dulation Generator · Flatness (A) · Digitizing · HP/Agilen t 54124T/ · Oscilloscop e · HP/Agilen t 54750A/ · HP/Agilen t 54752A · 40 mV p-p · 5 · x delay setting · Oscilloscop e · Prob es · Agilent 83630B
Page 49 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 49 · click to enlarge
Page 49

T able 1-3. Required Equipment for 83630B (continued)

Instrument Critical Sp eci cations Recommended

Division Ratio10:1 HP/Agilent 10431A Swept F requency Accuracy (P)

F requency Switching Time (P)

Pulse Mo dulation Video

Mo dulator O set and Gain (A)

Pulse Width 50 ns

Pulse P erformance (P)

Rise Time 10 ns

Pulse P erformance-Alt (P)

F requency10 Hz to 5 MHz

Pulse Mo dulation Video

F requency Accuracy 5 x 10

HP/Agilent 3325A/B AM Accuracy (P)

AM Dynamic Range (P)

100 kHz to 1 MHz 0.1 dB
100 kHz to 20 MHz 0.4 dB

Maxim um FM Deviation (P)

Mo dulation Meter (P)

Amplitude Range> 16 V p-p HP/Agilent 8111A/

Maxim um FM Deviation (P)

P o wer Meter P o wer Range: 1 W to 100 mW

P o wer Accuracy (P)

Accuracy 0.02 dB

AM Dynamic Range (P)

ALC P o wer Level

P o wer Meter P o wer Range: 1 W to 100 mW

HP/Agilent 437B P o wer Flatness (P)

Accuracy 0.02 dB

P o wer Flatness (A)

P o wer Meter P o wer Range: 1 W to 100 mW

HP/Agilent 438A P o wer Flatness and

Accuracy 0.02 dB

P o wer Flatness (AA)

Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AP)

1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Show 33 more labels / short text on this page
Use · Mo del · Oscilloscop e · Prob es · F eedthrough (P) · Amp/Mult Adjustments (A) · Amp/Filt Adjustments (A) · Pulse · HP/Agilent 8112B/ · Generator · HP/Agilent 8116A · F eedthrough (P) · F unction · 6 · Generator · Amplitude Accuracy · FM Accuracy (P) · AM Accuracy (A) · AM Delay (A) · FM Gain (A) · F unction · Generator · HP/Agilent 8116A/ · HP/Agilent 8175A · HP/Agilent 436A/ · HP/Agilent 437B/ · HP/Agilent 438A · Amp/Filt Adjustments (A) · Accuracy (A) · AM Accuracy (A) · Accuracy (AP) · 1-18 Required Equipment · Agilent 83630B
Page 50 of the Agilent Technologies 83640B user manual — AM
Page 50 · click to enlarge
Page 50

T able 1-3. Required Equipment for 83630B (continued)

Instrument Critical Sp eci cations Recommended

P o wer Sensor F requency Range: 10 MHz to 2.3 GHz

HP/Agilen t 8482A P o wer Flatness (P)

P o wer Range1 W to 100 mW

P o wer Flatness (A)

Step A ttenuator Flatness (AP)

P o wer Flatness and

Step A ttenuator Flatness (AA)

P o wer Flatness (AA)

P o wer Sensor F requency Range: 50 MHz to 26.5 GHz

HP/Agilen t 8485D Step A ttenuator Flatness (AA)

P o wer Range100 pW to 10 W

Step A ttenuator Flatness (AP)

P o wer Sensor F requency Range: 50 MHz to 26.5 GHz

HP/Agilen t 8485A P o wer Accuracy (P)

P o wer Range1 W to 100 mW

P o wer Flatness (P)

AM Dynamic Range (P)

ALC P o wer Level

P o wer Flatness (A)

P o wer Flatness and

P o wer Flatness (AA)

Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AP)

A ttenuator A ttenuation: 30 dB 0.05 dB at 50 MHz HP/Agilen t 11708A Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AP)

F requency Range (tuned):

2.5 MHz to 1.3 GHz

Range0 dBm to 127 dBm
Relative P o wer Accuracy 0.5 dB

AM

Rates20 Hz to 100 kHz
Depthto 99%
Accuracy 1% of reading 1 count

F requency Range (carrier):

HP/Agilen t 3048A Single Sideband Phase

0.01 to 18 GHz

Sensitivit y:

< 70 dBc at 100 Hz o set

< 78 dBc at 1 kHz o set

< 86 dBc at 10 kHz o set

< 107 dBc at 100 kHz o set

O set F requency Range:

100 Hz to 2 MHz

Amplitude Accuracy:

2 dB to 1 MHz o set

1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Agilent 8360 Required Equipment

Show 16 more labels / short text on this page
Use · Mo del · Accuracy (AP) · Amp/Filt Adjustments (A) · Accuracy (A) · AM Accuracy (A) · Accuracy (AP) · Measuring · HP/Agilen t 8902A · AM Accuracy (P) · Receiv er · Phase Noise · Measurement · Noise (P) · System · Agilent 83630B
Page 51 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 51 · click to enlarge
Page 51

T able 1-3. Required Equipment for 83630B (continued)

Instrument Critical Sp eci cations Recommended

F requency10 MHz
In ternal TimebaseAging
Stability< 1 x 10

10 MHz Standard (A)

F requency Range1.5 to 26.5 GHz

HP/Agilent 8348B FM Bandwidth (P)

Leveled Output P o wer 16 dBm
F requency Range100 kHz to 1.3 GHz

HP 8447F Pulse P erformance-Alt (P)

Preampli er Gain25 dB

Pulse Mo dulation Video

P o wer Ampli er Gain: 22 dB

1 meter of coax or

Mixer F requency Range1 GHz to 26.5 GHz

P/N 0955-0307 Pulse P erformance-Alt (P)

P o wer Splitter F requency Range: 10 MHz to 26.5 GHz HP/Agilent 11667B FM Bandwidth (P)

F requency RangeDC to 26.5 GHz

HP/Agilent 33330D External Leveling (P)

Maxim um Input200 mW
P olarit yNegativ e
A ttenuator F requency Range10 MHz to 26.5 GHz

Pulse P erformance-Alt (P)

Maxim um Input P o wer: 300 mW

Pulse Mo dulation Video

A ttenuation6 dB
A ttenuator F requency Range10 MHz to 26.5 GHz

P o wer Accuracy (P)

Maxim um Input P o wer: 300 mW

A ttenuation10 dB

Pulse P erformance (P)

Pulse P erformance-Alt (P)

FM Bandwidth (P) (2)

Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AA)

1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est

P { Manual P erformance T est

Show 38 more labels / short text on this page
Use · Mo del · F requency · HP/Agilent 5061A/ · 10 · Standard · /yr · HP/Agilent 5071A · Rate (P) · Microwa v e · Ampli er · Preampli er/ · P o wer · Ampli er · F eedthrough (P) · Delay Line · FM Bandwidth (P) · Discriminator · semi-rigid cable · (Norda 4817) · AM Accuracy (P) · FM Bandwidth (P) · Crystal · Detector · AM Bandwidth (P) · Amp/Mult Adjustments (A) · HP/Agilent 8493C · Opt 006 · F eedthrough (P) · HP/Agilent 8493C · Opt 010 · Spurious Signals · (Harmonic) (P) · AM Accuracy (P) · Amp/Mult Adjustments (A) · Amp/Filt Adjustments (A) · 1-20 Required Equipment · Agilent 83630B
Page 52 of the Agilent Technologies 83640B user manual — SLP-550
Page 52 · click to enlarge
Page 52

T able 1-3. Required Equipment for 83630B (continued)

Instrument Critical Sp eci cations Recommended

A ttenuator F requency Range10 MHz to 26.5 GHz

Maxim um Leveled P o wer (P)

Maxim um Input P o wer: 300 mW

Pulse P erformance (P)

A ttenuation20 dB

3.7 GHz Lo w

P/N 9135-0191 Pulse Mo dulation Video

K & L Microwa v e

Pulse Mo dulation Video

5LL30-130/BT2400/BP

500 MHz Lo w

Pulse Mo dulation Video

SLP-550

T o ol Kit No Substitute P/N 08360{60060

In v ertron California

Capacitor 1000 pf P/N 0160-4574 External Leveling (P)

F requency Range2 to 8 GHz

HP/Agilen t 11975A Spurious Signals

Leveled Output P o wer+16 dBm

Pulse P erformance-Alt (P)

F requency Range20 to 26.5 GHz HP/Agilen t 11970K Spurious Signals

Pulse P erformance-Alt (P)

WR42 to APC 3.5 mm (f ) K281C Spurious Signals

Pulse P erformance-Alt (P)

1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est

P { Manual P erformance T est

Agilent 8360 Required Equipment

Show 26 more labels / short text on this page
Use · Mo del · HP/Agilen t 8493C · Option 020 · P ass Filter · F eedthrough (P) · 130 MHz · Bessel Lo w · F eedthrough (P) · P ass Filter · Mini-Circuits · P ass Filter · F eedthrough (P) · Spurious Signals · Instruments 501TC · (Line Related) (P) · Microwa v e · Ampli er · (Harmonic) (P) · Harmonic · Mixer · (Harmonic) (P) · K{Band Coax · Adapter · (Harmonic) (P) · Agilent 83630B
Page 53 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 53 · click to enlarge
Page 53

The following list of adapters and cables is provided for con v enience. They may b e used in

equipmen t setups for p erformance tests or adjustmen ts.

SMA (m) to SMA (m) adapter 1250-1159

SMA (f ) to SMA (f ) adapter 1250-1158

SMB (m) to SMB (m) adapter 1250-0669

SMB (f ) to SMB (f ) adapter 1250-0672

SMB tee (f ) (m) (m) 1250-1391

3.5 mm (f ) to 3.5 mm (f ) adapter 5061-5311

3.5 mm (f ) to N-t yp e (m) adapter 1250-1745

2.4 mm (f ) to 2.92 mm (f ) adapter 1250-2187

2.4 mm (f ) to 2.4 mm (f ) adapter 1250-2188

2.4 mm (m) to 3.5 mm (f ) adapter 11901C

BNC (f ) to BNC (f ) adapter 1250-0080

BNC (m) to BNC (m) adapter 1250-0216

BNC (f ) to SMA (m) adapter 1250-1200

BNC (f ) to SMB (m) adapter 1250-1237

BNC tee (m) (f ) (f ) 1250-0781

SMB (f ) to BNC (m) exible cable 85680-60093

SMA semi-rigid cable 2 feet 08340-20124

GPIB cable 2 meters HP/Agilent 10833B

BNC cable 8120-2582 · 1-22 Required Equipment · Agilent 83630B

Page 54 of the Agilent Technologies 83640B user manual — ADC (AA)
Page 54 · click to enlarge
Page 54

Agilent 83640B Required Equipment

T able 1-4. Required Equipment for 83640B

Instrument Critical Sp eci cations Recommended

F requency Range10 MHz to 22 GHz

(40 GHz with external mixers)

F requency Span0 Hz, 100 Hz to 22 GHz
Amplitude Range+20 to 100 dBm
Resolution Bandwidth10 Hz to 3 MHz
Video Bandwidth10 Hz to 3 MHz
Log Fidelit y 0.1 dB/dB o v er 0 to

80 dB display (1.0 dB maxim um)

Video OutputDC v oltage prop ortional

Pulse Mo dulation On/O

to v ertical p osition of trace on display

Capable of phase lo cking to external

Pulse P erformance-Alt (P)

AM Dynamic Range (P)

Maxim um FM Deviation (P)

and API Spurs (A)

Square W a v e Symmetry (A)

A vs B Sweep Mo de

V ertical Sensitivit y5 mV/Div
Bandwidth100 MHz

Pulse P erformance-Alt (P)

F requency Range20 Hz to 10 MHz HP/Agilen t 3585A/B AM Bandwidth (P)

Controller 4 Mb yte RAM

Step A ttenuator Flatness (AP)

BASIC 5.1

P o wer Flatness and

GPIB

Step A ttenuator Flatness (AA)

ADC (AA)

P o wer Flatness (AA)

1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Agilent 8360 Required Equipment

Show 44 more labels / short text on this page
Use · Mo del · Sp ectrum · HP/Agilen t 8566B · Swept F requency · Analyzer · Accuracy (P) · Spurious Signals · (Harmonic) (P) · Spurious Signals · (Non-harmonic) (P) · Spurious Signals · (Line Related) (P) · Ratio (P) · 10 MHz reference · AM Accuracy (P) · FM Accuracy (P) · F ractional{N Reference · FM Gain (A) · Analog · HP/Agilen t 1740A/ · Amp/Mult Adjustments (A) · Oscilloscop e · HP/Agilen t 54600B · Amp/Filt Adjustments (A) · Lo cal · HP/Agilen t 83620/ · Oscillator · AM Accuracy (P) · HP/Agilen t 8340A/B · (Synthesized · Sweep er) · Sp ectrum · Analyzer with · FM Bandwidth (P) · T rac king · Generator · HP/Agilen t 9836/ · HP/Agilen t 9920/ · HP/Agilen t 310/ · Accuracy (AP) · HP/Agilen t 320 · YO Delay (AA) · Agilent 83640B
Page 55 of the Agilent Technologies 83640B user manual — ADC (AA)
Page 55 · click to enlarge
Page 55

T able 1-4. Required Equipment for 83640B (continued)

Instrument Critical Sp eci cations Recommended

P/N 08360-10001 Step A ttenuator Flatness (AP)

(Shipp ed with instrument)

P o wer Flatness and

Step A ttenuator Flatness (AA)

ADC (AA)

P o wer Flatness (AA)

D VM Range 50 to +50 VDC
Accuracy 0.01%
Input Imp edance 10 M

AM Dynamic Range (P)

Mo dulation Meter (P)

Lo w P o wer SRD Bias (A)

ADC (AA)

In ternal TimebaseAging
BandwidthDC to 300 MHz
Input Imp edance1 M and 50
V ertical Sensitivit y 5 mV/Div
Horizon tal Sensitivit y50 ns/Div

F requency Switching Time (P)

T riggerEv ent T riggerable

Pulse P erformance-Alt (P)

Pulse Mo dulation Video

In ternal Pulse Accuracy (P)

10 MHz Standard (A)

Mo dulator O set and Gain (A)

3 dB Bandwidth40 GHz

Pulse P erformance (P)

DC V oltage Accuracy 0.4% of full

scale 2 mV

T rigger Sensitivit y DC to 100 MHz:

T rigger & Timebase Jitter:

2.5 ps + 5E

Time In terv al Accuracy:

10 ps 0.1% of reading

Channel Input Re ection:

5% for 30 ps risetime

Division Ratio1:1 HP/Agilent 10437A In ternal Pulse Accuracy (P)

1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Show 42 more labels / short text on this page
Use · Mo del · Software No Substitute · Accuracy (AP) · YO Delay (AA) · HP/Agilent 3456A · / · External Leveling (P) · HP/Agilent 3457A · AM Accuracy (P) · AM Bandwidth (P) · FM Bandwidth (P) · AM Accuracy (A) · AM Delay (A) · Digitizing · Dual Channel · HP/Agilent 54110A/ · Oscilloscop e · Rate (P) · HP/Agilent 54111D · / · HP/Agilent 54222A/ · Swept F requency · HP/Agilent 54750A/ · Accuracy (P) · HP/Agilent 54752A · F eedthrough (P) · FM Bandwidth (P) · Mo dulation Generator · Flatness (A) · Digitizing · HP/Agilent 54124T/ · Oscilloscop e · HP/Agilent 54750A/ · HP/Agilent 54752A · 40 mV p-p · 5 · x delay setting · Oscilloscop e · Prob es · 1-24 Required Equipment · Agilent 83640B
Page 56 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 56 · click to enlarge
Page 56

T able 1-4. Required Equipment for 83640B (continued)

Instrument Critical Sp eci cations Recommended

Division Ratio10:1 HP/Agilen t 10431A Swept F requency

F requency Switc hing Time (P)

Pulse P erformance-Alt (P)

Pulse Mo dulation Video

Mo dulator O set and Gain (A)

Pulse Width 50 ns

Pulse P erformance (P)

Rise Time 10 ns

Pulse P erformance-Alt (P)

F requency10 Hz to 5 MHz

Pulse Mo dulation Video

F requency Accuracy 5 x 10

HP/Agilen t 3325A/B AM Accuracy (P)

AM Dynamic Range (P)

100 kHz to 1 MHz 0.1 dB
100 kHz to 20 MHz 0.4 dB

Maxim um FM Deviation (P)

Mo dulation Meter (P)

Amplitude Range> 16 V p-p HP/Agilen t 8111A/

Maxim um FM Deviation (P)

P o wer Meter P o wer Range: 1 W to 100 mW

P o wer Accuracy (P)

Accuracy 0.02 dB

AM Dynamic Range (P)

ALC P o wer Level

P o wer Meter P o wer Range: 1 W to 100 mW

HP/Agilen t 437B P o wer Flatness (P)

Accuracy 0.02 dB

P o wer Flatness (A)

P o wer Meter P o wer Range: 1 W to 100 mW

HP/Agilen t 438A P o wer Flatness and

Accuracy 0.02 dB

Step A ttenuator Flatness (AP)

P o wer Flatness (AA)

Step A ttenuator Flatness (AA)

1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Agilent 8360 Required Equipment

Show 33 more labels / short text on this page
Use · Mo del · Oscilloscop e · Prob es · Accuracy (P) · F eedthrough (P) · Amp/Mult Adjustments (A) · Amp/Filt Adjustments (A) · Pulse · HP/Agilen t 8112B/ · Generator · HP/Agilen t 8116A · F eedthrough (P) · 6 · F unction · Generator · Amplitude Accuracy · FM Accuracy (P) · AM Accuracy (A) · AM Delay (A) · FM Gain (A) · F unction · Generator · HP/Agilen t 8116A/ · HP/Agilen t 8175A · HP/Agilen t 436A/ · HP/Agilen t 437B/ · HP/Agilen t 438A · Amp/Filt Adjustments (A) · Accuracy (A) · AM Accuracy (A) · Accuracy (AP) · Agilent 83640B
Page 57 of the Agilent Technologies 83640B user manual — AM
Page 57 · click to enlarge
Page 57

T able 1-4. Required Equipment for 83640B (continued)

Instrument Critical Sp eci cations Recommended

P o wer Sensor F requency Range: 10 MHz to 2.3 GHz

HP/Agilent 8482A P o wer Flatness (P)

P o wer Range1 W to 100 mW

P o wer Flatness (A)

P o wer Flatness and

Step A ttenuator Flatness (AA)

P o wer Flatness (AA)

Step A ttenuator Flatness (AP)

P o wer Sensor F requency Range: 50 MHz to 40 GHz

HP/Agilent 8487A P o wer Accuracy (P)

P o wer Range1 W to 100 mW

P o wer Flatness (P)

AM Dynamic Range (P)

ALC P o wer Level

P o wer Flatness (A)

Square W a v e Symmetry (A)

Step A ttenuator Flatness (AP)

P o wer Flatness (AA)

Step A ttenuator Flatness (AA)

P o wer Sensor F requency Range: 50 MHz to 40 GHz

HP/Agilent 8487D Step A ttenuator Flatness (AA)

P o wer Range100 pW to 10 W

Step A ttenuator Flatness (AP)

A ttenuator A ttenuation: 30 dB 0.05 dB at 50 MHz HP/Agilent 11708A Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AP)

F requency Range (tuned):

2.5 MHz to 1.3 GHz

Step A ttenuator Flatness (AA)

Range0 dBm to 127 dBm

Step A ttenuator Flatness (AP)

Relative P o wer Accuracy 0.5 dB

AM

Rates20 Hz to 100 kHz
Depthto 99%
Accuracy 1% of reading 1 count

F requency Range (carrier):

HP/Agilent 3048A Single Sideband Phase

0.01 to 18 GHz

Sensitivit y:

< 70 dBc at 100 Hz o set

< 78 dBc at 1 kHz o set

< 86 dBc at 10 kHz o set

< 107 dBc at 100 kHz o set

O set F requency Range:

100 Hz to 2 MHz

Amplitude Accuracy:

2 dB to 1 MHz o set

1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Show 17 more labels / short text on this page
Use · Mo del · Accuracy (AP) · Amp/Filt Adjustments (A) · Accuracy (A) · AM Accuracy (A) · Accuracy (AP) · Measuring · HP/Agilent 8902A · AM Accuracy (P) · Receiver · Phase Noise · Measuremen t · Noise (P) · System · 1-26 Required Equipment · Agilent 83640B
Page 58 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 58 · click to enlarge
Page 58

T able 1-4. Required Equipment for 83640B (continued)

Instrument Critical Sp eci cations Recommended

F requency10 MHz
In ternal TimebaseAging
Stability< 1 x 10

10 MHz Standard (A)

F requency Range1.5 to 26.5 GHz

HP/Agilen t 8348A FM Bandwidth (P)

Leveled Output P o wer 16 dBm
F requency Range100 kHz to 1.3 GHz

HP/Agilen t 8447F Pulse P erformance-Alt (P)

Preampli er Gain25 dB

Pulse Mo dulation Video

P o wer Ampli er Gain: 22 dB

1 meter of coax or

Mixer F requency Range1 GHz to 20 GHz

P/N 0955-0307 Pulse P erformance-Alt (P)

P o wer Splitter F requency Range: 10 MHz to 26.5 GHz HP/Agilen t 11667C FM Bandwidth (P)

F requency RangeDC to 40 GHz

HP/Agilen t 33330E External Leveling (P)

Maxim um Input200 mW
P olarit yNegativ e
A ttenuator F requency Range10 MHz to 40 GHz

Pulse P erformance-Alt (P)

Maxim um Input P o wer: 300 mW

Pulse Mo dulation Video

A ttenuation6 dB
A ttenuator F requency Range10 MHz to 40 GHz

Maxim um Leveled P o wer (P)

Maxim um Input P o wer: 300 mW

Pulse P erformance (P)

A ttenuation20 dB
A ttenuator F requency Range50 MHz to 40 GHz

P o wer Accuracy (P)

Maxim um Input P o wer: 300 mW

A ttenuation20 dB

Pulse P erformance (P)

Pulse P erformance-Alt (P)

FM Bandwidth (P) (2)

Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AA)

1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est

P { Manual P erformance T est

Agilent 8360 Required Equipment

Show 39 more labels / short text on this page
Use · Mo del · F requency · HP/Agilen t 5061A/ · 10 · Standard · /yr · HP/Agilen t 5071A · Rate (P) · Microwa v e · Ampli er · Preampli er/ · P o wer · Ampli er · F eedthrough (P) · Delay Line · FM Bandwidth (P) · Discriminator · semi-rigid cable · (Norda 4817) · AM Accuracy (P) · FM Bandwidth (P) · Crystal · Detector · AM Bandwidth (P) · Amp/Mult Adjustments (A) · HP/Agilen t 8493C · Opt 006 · F eedthrough (P) · HP/Agilen t 8490D · Option 020 · HP/Agilen t 8490D · Option 010 · Spurious Signals · (Harmonic) (P) · AM Accuracy (P) · Amp/Mult Adjustments (A) · Amp/Filt Adjustments (A) · Agilent 83640B
Page 59 of the Agilent Technologies 83640B user manual — SLP-550
Page 59 · click to enlarge
Page 59

T able 1-4. Required Equipment for 83640B (continued)

Instrument Critical Sp eci cations Recommended

3.7 GHz Lo w

P/N 9135-0191 Pulse Mo dulation Video

K & L Microwa v e

Pulse Mo dulation Video

5LL30-130/BT2400/BP

500 MHz Lo w

Pulse Mo dulation Video

SLP-550

T o ol Kit No Substitute P/N 08360{60060

In v ertron California

Capacitor 1000 pf P/N 0160-4574 External Leveling (P)

F requency Range2 to 8 GHz

HP/Agilent 11975A Spurious Signals

Leveled Output P o wer+16 dBm

Pulse P erformance-Alt (P)

Mixer F requency Range26.5 to 40 GHz HP/Agilent 11970A Spurious Signals

Pulse P erformance-Alt (P)

Mixer F requency Range20 to 26.5 GHz HP/Agilent 11970K Spurious Signals

Pulse P erformance-Alt (P)

WR42 to APC 3.5 mm (f ) K281C Spurious Signals

Pulse P erformance-Alt (P)

WR28 to APC 2.4 mm (f ) R281A Spurious Signals

Pulse P erformance-Alt (P)

1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est

P { Manual P erformance T est

Show 28 more labels / short text on this page
Use · Mo del · P ass Filter · F eedthrough (P) · 130 MHz · Bessel Lo w · F eedthrough (P) · P ass Filter · Mini-Circuits · P ass Filter · F eedthrough (P) · Spurious Signals · Instruments 501TC · (Line Related) (P) · Microwa v e · Ampli er · (Harmonic) (P) · AM Accuracy (P) · (Harmonic) (P) · (Harmonic) (P) · K{Band Coax · Adapter · (Harmonic) (P) · R{Band Coax · Adapter · (Harmonic) (P) · 1-28 Required Equipment · Agilent 83640B
Page 60 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 60 · click to enlarge
Page 60

The follo wing list of adapters and cables is provided for con v enience. They may b e used in

equipmen t setups for p erformance tests or adjustmen ts.

SMA (m) to SMA (m) adapter 1250-1159

SMA (f ) to SMA (f ) adapter 1250-1158

SMB (m) to SMB (m) adapter 1250-0669

SMB (f ) to SMB (f ) adapter 1250-0672

SMB tee (f ) (m) (m) 1250-1391

3.5 mm (f ) to 3.5 mm (f ) adapter 5061-5311

3.5 mm (f ) to N-t yp e (m) adapter 1250-1745

2.4 mm (f ) to 2.92 mm (f ) adapter 1250-2187

2.4 mm (f ) to 2.4 mm (f ) adapter 1250-2188

2.4 mm (m) to 3.5 mm (f ) adapter 11901C

BNC (f ) to BNC (f ) adapter 1250-0080

BNC (m) to BNC (m) adapter 1250-0216

BNC (f ) to SMA (m) adapter 1250-1200

BNC (f ) to SMB (m) adapter 1250-1237

BNC tee (m) (f ) (f ) 1250-0781

SMB (f ) to BNC (m) exible cable 85680-60093

SMA semi-rigid cable 2 feet 08340-20124

GPIB cable 2 meters HP/Agilent 10833B

Agilent 8360 Required Equipment

BNC cable 8120-2582 · Agilent 83640B

Page 61 of the Agilent Technologies 83640B user manual — ADC (AA)
Page 61 · click to enlarge
Page 61

Agilent 83650B Required Equipment

T able 1-5. Required Equipment for 83650B

Instrument Critical Sp eci cations Recommended

F requency Range10 MHz to 22 GHz

(50 GHz with external mixers)

F requency Span0 Hz, 100 Hz to 22 GHz
Amplitude Range+20 to 100 dBm
Resolution Bandwidth10 Hz to 3 MHz
Video Bandwidth10 Hz to 3 MHz
Log Fidelity 0.1 dB/dB o v er 0 to

80 dB display (1.0 dB maxim um)

Video OutputDC v oltage prop ortional

Pulse Mo dulation On/O

to v ertical p osition of trace on display

Capable of phase lo cking to external

Pulse P erformance-Alt (P)

AM Dynamic Range (P)

Maxim um FM Deviation (P)

and API Spurs (A)

Square W a v e Symmetry (A)

A vs B Sweep Mo de

V ertical Sensitivit y5 mV/Div
Bandwidth100 MHz

Pulse P erformance-Alt (P)

F requency Range20 Hz to 10 MHz HP/Agilent 3585A/B AM Bandwidth (P)

Controller 4 Mb yte RAM

Step A ttenuator Flatness (AP)

BASIC 5.1

P o wer Flatness and

GPIB

Step A ttenuator Flatness (AA)

ADC (AA)

P o wer Flatness (AA)

1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Show 45 more labels / short text on this page
Use · Mo del · Sp ectrum · HP/Agilent 8566B · Swept F requency · Analyzer · Accuracy (P) · Spurious Signals · (Harmonic) (P) · Spurious Signals · (Non-harmonic) (P) · Spurious Signals · (Line Related) (P) · Ratio (P) · 10 MHz reference · AM Accuracy (P) · FM Accuracy (P) · F ractional{N Reference · FM Gain (A) · Analog · HP/Agilent 1740A/ · Amp/Mult Adjustments (A) · Oscilloscop e · HP/Agilent 54600B · Amp/Filt Adjustments (A) · Lo cal · HP/Agilent 83620/ · Oscillator · AM Accuracy (P) · HP/Agilent 8340A/B · (Synthesized · Sweep er) · Sp ectrum · Analyzer with · FM Bandwidth (P) · T rac king · Generator · HP/Agilent 9836/ · HP/Agilent 9920/ · HP/Agilent 310/ · Accuracy (AP) · HP/Agilent 320 · YO Delay (AA) · 1-30 Required Equipment · Agilent 83650B
Page 62 of the Agilent Technologies 83640B user manual — ADC (AA)
Page 62 · click to enlarge
Page 62

T able 1-5. Required Equipment for 83650B (continued)

Instrument Critical Sp eci cations Recommended

P/N 08360-10001 Step A ttenuator Flatness (AP)

(Shipp ed with instrument)

P o wer Flatness and

Step A ttenuator Flatness (AA)

ADC (AA)

P o wer Flatness (AA)

D VM Range 50 to +50 VDC
Accuracy 0.01%
Input Imp edance 10 M

AM Dynamic Range (P)

Mo dulation Meter (P)

Lo w P o wer SRD Bias (A)

ADC (AA)

In ternal TimebaseAging
BandwidthDC to 300 MHz
Input Imp edance1 M and 50
V ertical Sensitivit y 5 mV/Div
Horizon tal Sensitivit y50 ns/Div

F requency Switc hing Time (P)

T riggerEv ent T riggerable

Pulse P erformance-Alt (P)

Pulse Mo dulation Video

In ternal Pulse Accuracy (P)

10 MHz Standard (A)

Mo dulator O set and

3 dB Bandwidth50 GHz

Pulse P erformance (P)

DC V oltage Accuracy 0.4% of full

scale 2 mV

T rigger Sensitivit y DC to 100 MHz:

T rigger & Timebase Jitter:

2.5 ps + 5E

Time In terv al Accuracy:

10 ps 0.1% of reading

Channel Input Re ection:

5% for 30 ps risetime

Division Ratio1:1 HP/Agilen t 10437A In ternal Pulse Accuracy (P)

1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Agilent 8360 Required Equipment

Show 42 more labels / short text on this page
Use · Mo del · Software No Substitute · Accuracy (AP) · YO Delay (AA) · HP/Agilen t 3456A · / · External Leveling (P) · HP/Agilen t 3457A · AM Accuracy (P) · AM Bandwidth (P) · FM Bandwidth (P) · AM Accuracy (A) · AM Delay (A) · Digitizing · Dual Channel · HP/Agilen t 54110A/ · Oscilloscop e · Rate (P) · HP/Agilen t 54111D · / · HP/Agilen t 54222A/ · Swept F requency · HP/Agilen t 54750A/ · Accuracy (P) · HP/Agilen t 54752A · F eedthrough (P) · FM Bandwidth (P) · Gain (A) · Mo dulation Generator · Flatness (A) · Digitizing · HP/Agilen t 54124T/ · Oscilloscop e · HP/Agilen t 54750A/ · HP/Agilen t 54752A · 40 mV p-p · 5 · x delay setting · Oscilloscop e · Prob es · Agilent 83650B
Page 63 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 63 · click to enlarge
Page 63

T able 1-5. Required Equipment for 83650B (continued)

Instrument Critical Sp eci cations Recommended

Division Ratio10:1 HP/Agilent 10431A Swept F requency

F requency Switching Time (P)

Pulse P erformance-Alt (P)

Pulse Mo dulation Video

Mo dulator O set and

Pulse Width 50 ns

Pulse P erformance (P)

Rise Time 10 ns

Pulse P erformance-Alt (P)

F requency10 Hz to 5 MHz

Pulse Mo dulation Video

F requency Accuracy 5 x 10

HP/Agilent 3325A/B AM Accuracy (P)

AM Dynamic Range (P)

100 kHz to 1 MHz 0.1 dB
100 kHz to 20 MHz 0.4 dB

Maxim um FM Deviation (P)

Mo dulation Meter (P)

Amplitude Range> 16 V p-p HP/Agilent 8111A/

Maxim um FM Deviation (P)

P o wer Meter P o wer Range: 1 W to 100 mW

P o wer Accuracy (P)

Accuracy 0.02 dB

AM Dynamic Range (P)

ALC P o wer Level

P o wer Meter P o wer Range: 1 W to 100 mW

HP/Agilent 437B P o wer Flatness (P)

Accuracy 0.02 dB

P o wer Flatness (A)

P o wer Meter P o wer Range: 1 W to 100 mW

HP/Agilent 438A P o wer Flatness and

Accuracy 0.02 dB

Step A ttenuator Flatness (AP)

P o wer Flatness (AA)

Step A ttenuator Flatness (AA)

1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Show 35 more labels / short text on this page
Use · Mo del · Oscilloscop e · Prob es · Accuracy (P) · F eedthrough (P) · Amp/Mult Adjustments (A) · Amp/Filt Adjustments (A) · Gain (A) · Pulse · HP/Agilent 8112B/ · Generator · HP/Agilent 8116A · F eedthrough (P) · F unction · 6 · Generator · Amplitude Accuracy · FM Accuracy (P) · AM Accuracy (A) · AM Delay (A) · FM Gain (A) · F unction · Generator · HP/Agilent 8116A/ · HP/Agilent 8175A · HP/Agilent 436A/ · HP/Agilent 437B/ · HP/Agilent 438A · Amp/Filt Adjustments (A) · Accuracy (A) · AM Accuracy (A) · Accuracy (AP) · 1-32 Required Equipment · Agilent 83650B
Page 64 of the Agilent Technologies 83640B user manual — AM
Page 64 · click to enlarge
Page 64

T able 1-5. Required Equipment for 83650B (continued)

Instrument Critical Sp eci cations Recommended

P o wer Sensor F requency Range: 10 MHz to 2.3 GHz

HP/Agilen t 8482A P o wer Flatness (P)

P o wer Range1 W to 100 mW

P o wer Flatness (A)

P o wer Flatness and

Step A ttenuator Flatness (AA)

P o wer Flatness (AA)

Step A ttenuator Flatness (AP)

P o wer Sensor F requency Range: 50 MHz to 50 GHz

HP/Agilen t 8487A P o wer Accuracy (P)

P o wer Range1 W to 100 mW

P o wer Flatness (P)

AM Dynamic Range (P)

ALC P o wer Level

P o wer Flatness (A)

P o wer Flatness and

Step A ttenuator Flatness (AP)

P o wer Flatness (AA)

Step A ttenuator Flatness (AA)

P o wer Sensor F requency Range: 50 MHz to 50 GHz

HP/Agilen t 8487D Step A ttenuator Flatness (AP)

P o wer Range100 pW to 10 W

Step A ttenuator Flatness (AA)

A ttenuator A ttenuation: 30 dB 0.05 dB at 50 MHz HP/Agilen t 11708A Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AP)

F requency Range (tuned):

2.5 MHz to 1.3 GHz

Step A ttenuator Flatness (AA)

Range0 dBm to 127 dBm

Step A ttenuator Flatness (AP)

Relative P o wer Accuracy 0.5 dB

AM

Rates20 Hz to 100 kHz
Depthto 99%
Accuracy 1% of reading 1 count

F requency Range (carrier):

HP/Agilen t 3048A Single Sideband Phase

0.01 to 18 GHz

Sensitivit y:

< 70 dBc at 100 Hz o set

< 78 dBc at 1 kHz o set

< 86 dBc at 10 kHz o set

< 107 dBc at 100 kHz o set

O set F requency Range:

100 Hz to 2 MHz

Amplitude Accuracy:

2 dB to 1 MHz o set

1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Agilent 8360 Required Equipment

Show 16 more labels / short text on this page
Use · Mo del · Accuracy (AP) · Amp/Filt Adjustments (A) · Accuracy (A) · AM Accuracy (A) · Accuracy (AP) · Measuring · HP/Agilen t 8902A · AM Accuracy (P) · Receiv er · Phase Noise · Measurement · Noise (P) · System · Agilent 83650B
Page 65 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 65 · click to enlarge
Page 65

T able 1-5. Required Equipment for 83650B (continued)

Instrument Critical Sp eci cations Recommended

F requency10 MHz
In ternal TimebaseAging
Stability< 1 x 10

10 MHz Standard (A)

F requency Range1.5 to 26.5 GHz

HP/Agilent 8348A FM Bandwidth (P)

Leveled Output P o wer 16 dBm
F requency Range100 kHz to 1.3 GHz

HP/Agilent 8447F Pulse P erformance-Alt (P)

Preampli er Gain25 dB

Pulse Mo dulation Video

P o wer Ampli er Gain: 22 dB

1 meter of coax or

Mixer F requency Range1 GHz to 26.5 GHz

P/N 0955-0307 Pulse P erformance-Alt (P)

P o wer Splitter F requency Range: 10 MHz to 50 GHz HP/Agilent 11667C FM Bandwidth (P)

F requency RangeDC to 50 GHz

HP/Agilent 33330E External Leveling (P)

Maxim um Input200 mW
P olarit yNegativ e
A ttenuator F requency Range10 MHz to 50 GHz

Pulse P erformance-Alt (P)

Maxim um Input P o wer: 300 mW

Pulse Mo dulation Video

A ttenuation6 dB
A ttenuator F requency Range10 MHz to 50 GHz

Maxim um Leveled P o wer (P)

Maxim um Input P o wer: 300 mW

Pulse P erformance (P)

A ttenuation20 dB
A ttenuator F requency Range50 MHz to 50 GHz

P o wer Accuracy (P)

Maxim um Input P o wer: 300 mW

A ttenuation20 dB

Pulse P erformance (P)

Pulse P erformance-Alt (P)

FM Bandwidth (P) (2)

Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AA)

1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est

P { Manual P erformance T est

Show 40 more labels / short text on this page
Use · Mo del · F requency · HP/Agilent 5061A/ · 10 · Standard · /yr · HP/Agilent 5071A · Rate (P) · Microwa v e · Ampli er · Preampli er/ · P o wer · Ampli er · F eedthrough (P) · Delay Line · FM Bandwidth (P) · Discriminator · semi-rigid cable · (Norda 4817) · AM Accuracy (P) · FM Bandwidth (P) · Crystal · Detector · AM Bandwidth (P) · Amp/Mult Adjustments (A) · HP/Agilent 8490D · Opt 006 · F eedthrough (P) · HP/Agilent 8490D · Option 020 · HP/Agilent 8490D · Option 010 · Spurious Signals · (Harmonic) (P) · AM Accuracy (P) · Amp/Mult Adjustments (A) · Amp/Filt Adjustments (A) · 1-34 Required Equipment · Agilent 83650B
Page 66 of the Agilent Technologies 83640B user manual — SLP-550
Page 66 · click to enlarge
Page 66

T able 1-5. Required Equipment for 83650B (continued)

Instrument Critical Sp eci cations Recommended

3.7 GHz Lo w

P/N 9135-0191 Pulse Mo dulation Video

K & L Microwa v e

Pulse Mo dulation Video

5LL30-130/BT2400/BP

500 MHz Lo w

Pulse Mo dulation Video

SLP-550

T o ol Kit No Substitute P/N 08360{60060

In v ertron California

Capacitor 1000 pf P/N 0160-4574 External Leveling (P)

Mixer F requency Range26.5 to 40 GHz HP/Agilen t 11970A Spurious Signals

Pulse P erformance-Alt (P)

F requency Range2 to 8 GHz

HP/Agilen t 11975A Spurious Signals

Leveled Output P o wer+16 dBm

Pulse P erformance-Alt (P)

Mixer F requency Range20 to 26.5 GHz HP/Agilen t 11970K Spurious Signals

Pulse P erformance-Alt (P)

Mixer F requency Range40 to 50 GHz HP/Agilen t 11970Q Spurious Signals

Pulse P erformance-Alt (P)

WR42 to APC 3.5 mm (f ) K281C Spurious Signals

Pulse P erformance-Alt (P)

1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est

P { Manual P erformance T est

Agilent 8360 Required Equipment

Show 25 more labels / short text on this page
Use · Mo del · P ass Filter · F eedthrough (P) · 130 MHz · Bessel Lo w · F eedthrough (P) · P ass Filter · Mini-Circuits · P ass Filter · F eedthrough (P) · Spurious Signals · Instruments 501TC · (Line Related) (P) · (Harmonic) (P) · Microwa v e · Ampli er · (Harmonic) (P) · AM Accuracy (P) · (Harmonic) (P) · (Harmonic) (P) · K{Band Coax · Adapter · (Harmonic) (P) · Agilent 83650B
Page 67 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 67 · click to enlarge
Page 67

T able 1-5. Required Equipment for 83650B (continued)

Instrument Critical Sp eci cations Recommended

WR28 to APC 2.4 mm (f ) R281A Spurious Signals

Pulse P erformance-Alt (P)

WR22 to APC 2.4 mm (f ) Q281A Spurious Signals

Pulse P erformance-Alt (P)

1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est

P { Manual P erformance T est

Show 10 more labels / short text on this page
Use · Mo del · R{Band Coax · Adapter · (Harmonic) (P) · Q{Band Coax · Adapter · (Harmonic) (P) · 1-36 Required Equipment · Agilent 83650B
Page 68 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 68 · click to enlarge
Page 68

The follo wing list of adapters and cables is provided for con v enience. They may b e used in

equipmen t setups for p erformance tests or adjustmen ts.

SMA (m) to SMA (m) adapter 1250-1159

SMA (f ) to SMA (f ) adapter 1250-1158

SMB (m) to SMB (m) adapter 1250-0669

SMB (f ) to SMB (f ) adapter 1250-0672

SMB tee (f ) (m) (m) 1250-1391

3.5 mm (f ) to 3.5 mm (f ) adapter 5061-5311

3.5 mm (f ) to N-t yp e (m) adapter 1250-1745

2.4 mm (f ) to 2.92 mm (f ) adapter 1250-2187

2.4 mm (f ) to 2.4 mm (f ) adapter 1250-2188

2.4 mm (m) to 3.5 mm (f ) adapter 11901C

BNC (f ) to BNC (f ) adapter 1250-0080

BNC (m) to BNC (m) adapter 1250-0216

BNC (f ) to SMA (m) adapter 1250-1200

BNC (f ) to SMB (m) adapter 1250-1237

BNC tee (m) (f ) (f ) 1250-0781

SMB (f ) to BNC (m) exible cable 85680-60093

SMA semi-rigid cable 2 feet 08340-20124

GPIB cable 2 meters HP/Agilent 10833B

Agilent 8360 Required Equipment

BNC cable 8120-2582 · Agilent 83650B

Page 69 of the Agilent Technologies 83640B user manual — ADC (AA)
Page 69 · click to enlarge
Page 69

Agilent 83623L Required Equipment

T able 1-6. Required Equipment for 83623L

Instrument Critical Sp eci cations Recommended

F requency Range10 MHz to 20 GHz
F requency Span0 Hz, 100 Hz to 20 GHz
Amplitude Range+20 to 100 dBm
Resolution Bandwidth10 Hz to 3 MHz
Video Bandwidth10 Hz to 3 MHz
Log Fidelity 0.1 dB/dB o v er 0 to

80 dB display (1.0 dB maxim um)

Video OutputDC v oltage prop ortional

to v ertical p osition of trace on display

Capable of phase lo cking to external

and API Spurs (A)

Square W a v e Symmetry (A)

A vs B Sweep Mo de

V ertical Sensitivit y5 mV/Div
Bandwidth100 MHz

AM O set (A)

Controller 4 Mb yte RAM

Step A ttenuator Flatness (AP)

BASIC 5.1

P o wer Flatness and

GPIB

Step A ttenuator Flatness (AA)

ADC (AA)

P o wer Flatness (AA)

P/N 08360-10001 Step A ttenuator Flatness (AP)

(Shipp ed with instrument)

P o wer Flatness and

Step A ttenuator Flatness (AA)

ADC (AA)

P o wer Flatness (AA)

D VM Range 50 to +50 VDC
Accuracy 0.01%

Lo w P o wer SRD Bias (A)

Input Imp edance 10 M

ADC (AA)

1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Show 36 more labels / short text on this page
Use · Mo del · Sp ectrum · HP/Agilent 8566B · Swept F requency · Analyzer · Accuracy (P) · Spurious Signals · (Harmonic) (P) · Spurious Signals · (Non-harmonic) (P) · Spurious Signals · (Line Related) (P) · F ractional{N Reference · 10 MHz reference · Analog · HP/Agilent 1740A/ · Amp/Mult Adjustments (A) · Oscilloscop e · HP/Agilent 54600B · Amp/Filt Adjustments (A) · HP/Agilent 9836/ · HP/Agilent 9920/ · HP/Agilent 310/ · Accuracy (AP) · HP/Agilent 320 · YO Delay (AA) · Software No Substitute · Accuracy (AP) · YO Delay (AA) · HP/Agilent 3456A · / · External Leveling (P) · HP/Agilent 3457A · 1-38 Required Equipment · Agilent 83623L
Page 70 of the Agilent Technologies 83640B user manual — HP 54752A
Page 70 · click to enlarge
Page 70

T able 1-6. Required Equipment for 83623L (continued)

Instrument Critical Sp eci cations Recommended

In ternal TimebaseAging
BandwidthDC to 300 MHz
Input Imp edance1 M and 50
V ertical Sensitivit y 5 mV/Div

HP 54750A/

Horizon tal Sensitivit y50 ns/Div

HP 54752A

F requency Switc hing Time (P)

T riggerEv ent T riggerable

10 MHz Standard (A)

Division Ratio10:1 HP 10431A Swept F requency

F requency Switc hing Time (P)

P o wer Meter P o wer Range: 1 W to 100 mW

P o wer Accuracy (P)

Accuracy 0.02 dB

ALC P o wer Level

P o wer Meter P o wer Range: 1 W to 100 mW

HP/Agilen t 437B P o wer Flatness (P)

Accuracy 0.02 dB

P o wer Flatness (A)

P o wer Meter P o wer Range: 1 W to 100 mW

HP/Agilen t 438A P o wer Flatness and

Accuracy 0.02 dB

P o wer Flatness (AA)

Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AP)

1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Agilent 8360 Required Equipment

Show 24 more labels / short text on this page
Use · Mo del · Digitizing · Dual Channel · HP/Agilen t 54110A/ · Oscilloscop e · HP/Agilen t 54111D · / · Rate (P) · HP/Agilen t 54222A/ · Swept F requency · Accuracy (P) · Oscilloscop e · Prob es · Accuracy (P) · Amp/Mult Adjustments (A) · Amp/Filt Adjustments (A) · HP/Agilen t 436A/ · HP/Agilen t 437B/ · Amp/Filt Adjustments (A) · HP/Agilen t 438A · Accuracy (A) · Accuracy (AP) · Agilent 83623L
Page 71 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 71 · click to enlarge
Page 71

T able 1-6. Required Equipment for 83623L (continued)

Instrument Critical Sp eci cations Recommended

P o wer Sensor F requency Range: 10 MHz to 2.3 GHz

HP/Agilent 8482A P o wer Flatness (P)

P o wer Range1 W to 100 mW

P o wer Flatness (A)

Step A ttenuator Flatness (AP)

P o wer Flatness and

Step A ttenuator Flatness (AA)

P o wer Flatness (AA)

P o wer Sensor F requency Range: 50 MHz to 20 GHz

HP/Agilent 8485D Step A ttenuator Flatness (AA)

P o wer Range100 pW to 10 W

Step A ttenuator Flatness (AP)

P o wer Sensor F requency Range: 50 MHz to 20 GHz

HP/Agilent 8485A P o wer Accuracy (P)

P o wer Range1 W to 100 mW

P o wer Flatness (P)

ALC P o wer Level

P o wer Flatness (A)

P o wer Flatness and

P o wer Flatness (AA)

Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AP)

A ttenuator A ttenuation: 30 dB 0.05 dB at 50 MHz HP/Agilent 11708A Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AP)

F requency Range (carrier):

HP/Agilent 3048A Single Sideband Phase

0.01 to 18 GHz

Sensitivit y:

< 70 dBc at 100 Hz o set

< 78 dBc at 1 kHz o set

< 86 dBc at 10 kHz o set

< 107 dBc at 100 kHz o set

O set F requency Range:

100 Hz to 2 MHz

Amplitude Accuracy:

2 dB to 1 MHz o set

F requency10 MHz
In ternal TimebaseAging
Stability< 1 x 10

10 MHz Standard (A)

F requency RangeDC to 20 GHz

HP/Agilent 33330D External Leveling (P)

Maxim um Input200 mW

AM O set (A)

P olarit yNegativ e

1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est

P { Manual P erformance T est

Show 21 more labels / short text on this page
Use · Mo del · Accuracy (AP) · Amp/Filt Adjustments (A) · Accuracy (A) · Accuracy (AP) · Phase Noise · Measuremen t · Noise (P) · System · F requency · HP/Agilent 5061A/ · 10 · Standard · /yr · HP/Agilent 5071A · Rate (P) · Crystal · Detector · 1-40 Required Equipment · Agilent 83623L
Page 72 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 72 · click to enlarge
Page 72

T able 1-6. Required Equipment for 83623L (continued)

Instrument Critical Sp eci cations Recommended

A ttenuator F requency Range10 MHz to 20 GHz

P o wer Accuracy (P)

Maxim um Input P o wer: 300 mW

A ttenuation10 dB

Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AA)

A ttenuator F requency Range10 MHz to 20 GHz

Maxim um Leveled P o wer (P)

Maxim um Input P o wer: 300 mW

A ttenuation20 dB

T o ol Kit No Substitute P/N 08360{60060

In v ertron California

Capacitor 1000 pf P/N 0160-4574 External Leveling (P)

1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est

P { Manual P erformance T est

Agilent 8360 Required Equipment

Show 14 more labels / short text on this page
Use · Mo del · HP/Agilen t 8493C · Opt 010 · Spurious Signals · (Harmonic) (P) · Amp/Mult Adjustments (A) · Amp/Filt Adjustments (A) · HP/Agilen t 8493C · Option 020 · Spurious Signals · Instruments 501TC · (Line Related) (P) · Agilent 83623L
Page 73 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 73 · click to enlarge
Page 73

The following list of adapters and cables is provided for con v enience. They may b e used in

equipmen t setups for p erformance tests or adjustmen ts.

SMA (m) to SMA (m) adapter 1250-1159

SMA (f ) to SMA (f ) adapter 1250-1158

SMB (m) to SMB (m) adapter 1250-0669

SMB (f ) to SMB (f ) adapter 1250-0672

SMB tee (f ) (m) (m) 1250-1391

3.5 mm (f ) to 3.5 mm (f ) adapter 5061-5311

3.5 mm (f ) to N-t yp e (m) adapter 1250-1745

2.4 mm (f ) to 2.92 mm (f ) adapter 1250-2187

2.4 mm (f ) to 2.4 mm (f ) adapter 1250-2188

2.4 mm (m) to 3.5 mm (f ) adapter 11901C

BNC (f ) to BNC (f ) adapter 1250-0080

BNC (m) to BNC (m) adapter 1250-0216

BNC (f ) to SMA (m) adapter 1250-1200

BNC (f ) to SMB (m) adapter 1250-1237

BNC tee (m) (f ) (f ) 1250-0781

SMB (f ) to BNC (m) exible cable 85680-60093

SMA semi-rigid cable 2 feet 08340-20124

GPIB cable 2 meters HP/Agilent 10833B

BNC cable 8120-2582 · 1-42 Required Equipment · Agilent 83623L

Page 74 of the Agilent Technologies 83640B user manual — ADC (AA)
Page 74 · click to enlarge
Page 74

Agilent 83630L Required Equipment

T able 1-7. Required Equipment for 83630L

Instrument Critical Sp eci cations Recommended

F requency Range10 MHz to 22 GHz

(26.5 GHz with external mixers)

F requency Span0 Hz, 100 Hz to 22 GHz
Amplitude Range+20 to 100 dBm
Resolution Bandwidth10 Hz to 3 MHz
Video Bandwidth10 Hz to 3 MHz
Log Fidelit y 0.1 dB/dB o v er 0 to

80 dB display (1.0 dB maxim um)

Video OutputDC v oltage prop ortional

to v ertical p osition of trace on display

and API Spurs (A)

Capable of phase lo cking to external

Square W a v e Symmetry (A)

A vs B Sweep Mo de

V ertical Sensitivit y5 mV/Div
Bandwidth100 MHz

Controller 4 Mb yte RAM

Step A ttenuator Flatness (AP)

BASIC 5.1

P o wer Flatness and

GPIB

Step A ttenuator Flatness (AA)

ADC (AA)

P o wer Flatness (AA)

P/N 08360-10001 Step A ttenuator Flatness (AP)

(Shipp ed with instrument)

P o wer Flatness and

Step A ttenuator Flatness (AA)

ADC (AA)

P o wer Flatness (AA)

D VM Range 50 to +50 VDC
Accuracy 0.01%

Lo w P o wer SRD Bias (A)

Input Imp edance 10 M

ADC (AA)

1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Agilent 8360 Required Equipment

Show 35 more labels / short text on this page
Use · Mo del · Sp ectrum · HP/Agilen t 8566B · Swept F requency · Analyzer · Accuracy (P) · Spurious Signals · (Harmonic) (P) · Spurious Signals · (Non-harmonic) (P) · Spurious Signals · (Line Related) (P) · F ractional{N Reference · 10 MHz reference · Analog · HP/Agilen t 1740A/ · Amp/Mult Adjustments (A) · Oscilloscop e · HP/Agilen t 54600B · Amp/Filt Adjustments (A) · HP/Agilen t 9836/ · HP/Agilen t 9920/ · HP/Agilen t 310/ · Accuracy (AP) · HP/Agilen t 320 · YO Delay (AA) · Software No Substitute · Accuracy (AP) · YO Delay (AA) · HP/Agilen t 3456A · / · External Leveling (P) · HP/Agilen t 3457A · Agilent 83630L
Page 75 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 75 · click to enlarge
Page 75

T able 1-7. Required Equipment for 83630L (continued)

Instrument Critical Sp eci cations Recommended

In ternal TimebaseAging
BandwidthDC to 300 MHz
Input Imp edance1 M and 50
V ertical Sensitivit y 5 mV/Div
Horizon tal Sensitivit y50 ns/Div

F requency Switching Time (P)

T riggerEv ent T riggerable

10 MHz Standard (A)

Division Ratio10:1 HP/Agilent 10431A Swept F requency Accuracy (P)

F requency Switching Time (P)

AM O set (A)

P o wer Meter P o wer Range: 1 W to 100 mW

P o wer Accuracy (P)

Accuracy 0.02 dB

ALC P o wer Level

P o wer Meter P o wer Range: 1 W to 100 mW

HP/Agilent 437B P o wer Flatness (P)

Accuracy 0.02 dB

P o wer Flatness (A)

P o wer Meter P o wer Range: 1 W to 100 mW

HP/Agilent 438A P o wer Flatness and

Accuracy 0.02 dB

P o wer Flatness (AA)

Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AP)

P o wer Sensor F requency Range: 10 MHz to 2.3 GHz

HP/Agilent 8482A P o wer Flatness (P)

P o wer Range1 W to 100 mW

P o wer Flatness (A)

Step A ttenuator Flatness (AP)

P o wer Flatness and

Step A ttenuator Flatness (AA)

P o wer Flatness (AA)

1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Show 27 more labels / short text on this page
Use · Mo del · Digitizing · Dual Channel · HP/Agilent 54110A/ · Oscilloscop e · HP/Agilent 54111D · / · Rate (P) · HP/Agilent 54222A/ · Swept F requency · Accuracy (P) · HP/Agilent 54750A/ · HP/Agilent 54752A · Oscilloscop e · Prob es · Amp/Mult Adjustments (A) · Amp/Filt Adjustments (A) · HP/Agilent 436A/ · HP/Agilent 437B/ · Amp/Filt Adjustments (A) · HP/Agilent 438A · Accuracy (A) · Accuracy (AP) · Accuracy (AP) · 1-44 Required Equipment · Agilent 83630L
Page 76 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 76 · click to enlarge
Page 76

T able 1-7. Required Equipment for 83630L (continued)

Instrument Critical Sp eci cations Recommended

P o wer Sensor F requency Range: 50 MHz to 26.5 GHz

HP/Agilen t 8485D Step A ttenuator Flatness (AA)

P o wer Range100 pW to 10 W

Step A ttenuator Flatness (AP)

P o wer Sensor F requency Range: 50 MHz to 26.5 GHz

HP/Agilen t 8485A P o wer Accuracy (P)

P o wer Range1 W to 100 mW

P o wer Flatness (P)

ALC P o wer Level

P o wer Flatness (A)

P o wer Flatness and

P o wer Flatness (AA)

Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AP)

A ttenuator A ttenuation: 30 dB 0.05 dB at 50 MHz HP/Agilen t 11708A Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AP)

F requency Range (carrier):

HP/Agilen t 3048A Single Sideband Phase

0.01 to 18 GHz

Sensitivit y:

< 70 dBc at 100 Hz o set

< 78 dBc at 1 kHz o set

< 86 dBc at 10 kHz o set

< 107 dBc at 100 kHz o set

O set F requency Range:

100 Hz to 2 MHz

Amplitude Accuracy:

2 dB to 1 MHz o set

F requency10 MHz
In ternal TimebaseAging
Stability< 1 x 10

10 MHz Standard (A)

F requency RangeDC to 26.5 GHz

HP/Agilen t 33330D External Leveling (P)

Maxim um Input200 mW
P olarit yNegativ e

AM O set (A)

1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est

P { Manual P erformance T est

Agilent 8360 Required Equipment

Show 20 more labels / short text on this page
Use · Mo del · Amp/Filt Adjustments (A) · Accuracy (A) · Accuracy (AP) · Phase Noise · Measurement · Noise (P) · System · F requency · HP/Agilen t 5061A/ · Standard · 10 · /yr · HP/Agilen t 5071A · Rate (P) · Crystal · Detector · Amp/Mult Adjustments (A) · Agilent 83630L
Page 77 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 77 · click to enlarge
Page 77

T able 1-7. Required Equipment for 83630L (continued)

Instrument Critical Sp eci cations Recommended

A ttenuator F requency Range10 MHz to 26.5 GHz

P o wer Accuracy (P)

Maxim um Input P o wer: 300 mW

A ttenuation10 dB

Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AA)

A ttenuator F requency Range10 MHz to 26.5 GHz

Maxim um Leveled P o wer (P)

Maxim um Input P o wer: 300 mW

A ttenuation20 dB

T o ol Kit No Substitute P/N 08360{60060

In v ertron California

Capacitor 1000 pf P/N 0160-4574 External Leveling (P)

F requency Range2 to 8 GHz

HP/Agilent 11975A Spurious Signals

Leveled Output P o wer+16 dBm
F requency Range20 to 26.5 GHz HP/Agilent 11970K Spurious Signals

WR42 to APC 3.5 mm (f ) K281C Spurious Signals

1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est

P { Manual P erformance T est

Show 24 more labels / short text on this page
Use · Mo del · HP/Agilent 8493C · Opt 010 · Spurious Signals · (Harmonic) (P) · Amp/Mult Adjustments (A) · Amp/Filt Adjustments (A) · HP/Agilent 8493C · Option 020 · Spurious Signals · Instruments 501TC · (Line Related) (P) · Microwa v e · Ampli er · (Harmonic) (P) · Harmonic · Mixer · (Harmonic) (P) · K{Band Coax · Adapter · (Harmonic) (P) · 1-46 Required Equipment · Agilent 83630L
Page 78 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 78 · click to enlarge
Page 78

The follo wing list of adapters and cables is provided for con v enience. They may b e used in

equipmen t setups for p erformance tests or adjustmen ts.

SMA (m) to SMA (m) adapter 1250-1159

SMA (f ) to SMA (f ) adapter 1250-1158

SMB (m) to SMB (m) adapter 1250-0669

SMB (f ) to SMB (f ) adapter 1250-0672

SMB tee (f ) (m) (m) 1250-1391

3.5 mm (f ) to 3.5 mm (f ) adapter 5061-5311

3.5 mm (f ) to N-t yp e (m) adapter 1250-1745

2.4 mm (f ) to 2.92 mm (f ) adapter 1250-2187

2.4 mm (f ) to 2.4 mm (f ) adapter 1250-2188

2.4 mm (m) to 3.5 mm (f ) adapter 11901C

BNC (f ) to BNC (f ) adapter 1250-0080

BNC (m) to BNC (m) adapter 1250-0216

BNC (f ) to SMA (m) adapter 1250-1200

BNC (f ) to SMB (m) adapter 1250-1237

BNC tee (m) (f ) (f ) 1250-0781

SMB (f ) to BNC (m) exible cable 85680-60093

SMA semi-rigid cable 2 feet 08340-20124

GPIB cable 2 meters HP/Agilent 10833B

Agilent 8360 Required Equipment

BNC cable 8120-2582 · Agilent 83630L

Page 79 of the Agilent Technologies 83640B user manual — ADC (AA)
Page 79 · click to enlarge
Page 79

Agilent 83640L Required Equipment

T able 1-8. Required Equipment for 83640L

Instrument Critical Sp eci cations Recommended

F requency Range10 MHz to 22 GHz

(40 GHz with external mixers)

F requency Span0 Hz, 100 Hz to 22 GHz
Amplitude Range+20 to 100 dBm
Resolution Bandwidth10 Hz to 3 MHz
Video Bandwidth10 Hz to 3 MHz
Log Fidelity 0.1 dB/dB o v er 0 to

80 dB display (1.0 dB maxim um)

Video OutputDC v oltage prop ortional

to v ertical p osition of trace on display

and API Spurs (A)

Capable of phase lo cking to external

Square W a v e Symmetry (A)

A vs B Sweep Mo de

V ertical Sensitivit y5 mV/Div
Bandwidth100 MHz

AM O set (A)

Controller 4 Mb yte RAM

Step A ttenuator Flatness (AP)

BASIC 5.1

P o wer Flatness and

GPIB

Step A ttenuator Flatness (AA)

ADC (AA)

P o wer Flatness (AA)

P/N 08360-10001 Step A ttenuator Flatness (AP)

(Shipp ed with instrument)

P o wer Flatness and

Step A ttenuator Flatness (AA)

ADC (AA)

P o wer Flatness (AA)

D VM Range 50 to +50 VDC
Accuracy 0.01%

Lo w P o wer SRD Bias (A)

Input Imp edance 10 M

ADC (AA)

1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Show 36 more labels / short text on this page
Use · Mo del · Sp ectrum · HP/Agilent 8566B · Swept F requency · Analyzer · Accuracy (P) · Spurious Signals · (Harmonic) (P) · Spurious Signals · (Non-harmonic) (P) · Spurious Signals · (Line Related) (P) · F ractional{N Reference · 10 MHz reference · Analog · HP/Agilent 1740A/ · Amp/Mult Adjustments (A) · Oscilloscop e · HP/Agilent 54600B · Amp/Filt Adjustments (A) · HP/Agilent 9836/ · HP/Agilent 9920/ · HP/Agilent 310/ · Accuracy (AP) · HP/Agilent 320 · YO Delay (AA) · Software No Substitute · Accuracy (AP) · YO Delay (AA) · HP/Agilent 3456A · / · External Leveling (P) · HP/Agilent 3457A · 1-48 Required Equipment · Agilent 83640L
Page 80 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 80 · click to enlarge
Page 80

T able 1-8. Required Equipment for 83640L (continued)

Instrument Critical Sp eci cations Recommended

In ternal TimebaseAging
BandwidthDC to 300 MHz
Input Imp edance1 M and 50
V ertical Sensitivit y 5 mV/Div
Horizon tal Sensitivit y50 ns/Div

F requency Switc hing Time (P)

T riggerEv ent T riggerable

10 MHz Standard (A)

Division Ratio10:1 HP/Agilen t 10431A Swept F requency

F requency Switc hing Time (P)

P o wer Meter P o wer Range: 1 W to 100 mW

P o wer Accuracy (P)

Accuracy 0.02 dB

ALC P o wer Level

P o wer Meter P o wer Range: 1 W to 100 mW

HP/Agilen t 437B P o wer Flatness (P)

Accuracy 0.02 dB

P o wer Flatness (A)

P o wer Meter P o wer Range: 1 W to 100 mW

HP/Agilen t 438A P o wer Flatness and

Accuracy 0.02 dB

Step A ttenuator Flatness (AP)

P o wer Flatness (AA)

Step A ttenuator Flatness (AA)

P o wer Sensor F requency Range: 10 MHz to 2.3 GHz

HP/Agilen t 8482A P o wer Flatness (P)

P o wer Range1 W to 100 mW

P o wer Flatness (A)

P o wer Flatness and

Step A ttenuator Flatness (AA)

P o wer Flatness (AA)

Step A ttenuator Flatness (AP)

1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Agilent 8360 Required Equipment

Show 27 more labels / short text on this page
Use · Mo del · Digitizing · Dual Channel · HP/Agilen t 54110A/ · Oscilloscop e · HP/Agilen t 54111D · / · Rate (P) · HP/Agilen t 54222A/ · Swept F requency · Accuracy (P) · HP/Agilen t 54750A/ · HP/Agilen t 54752A · Oscilloscop e · Prob es · Accuracy (P) · Amp/Mult Adjustments (A) · Amp/Filt Adjustments (A) · HP/Agilen t 436A/ · HP/Agilen t 437B/ · Amp/Filt Adjustments (A) · HP/Agilen t 438A · Accuracy (A) · Accuracy (AP) · Accuracy (AP) · Agilent 83640L
Page 81 of the Agilent Technologies 83640B user manual — AM
Page 81 · click to enlarge
Page 81

T able 1-8. Required Equipment for 83640L (continued)

Instrument Critical Sp eci cations Recommended

P o wer Sensor F requency Range: 50 MHz to 40 GHz

HP/Agilent 8487A P o wer Accuracy (P)

P o wer Range1 W to 100 mW

P o wer Flatness (P)

ALC P o wer Level

P o wer Flatness (A)

Square W a v e Symmetry (A)

Step A ttenuator Flatness (AP)

P o wer Flatness (AA)

Step A ttenuator Flatness (AA)

P o wer Sensor F requency Range: 50 MHz to 40 GHz

HP/Agilent 8487D Step A ttenuator Flatness (AA)

P o wer Range100 pW to 10 W

Step A ttenuator Flatness (AP)

A ttenuator A ttenuation: 30 dB 0.05 dB at 50 MHz HP/Agilent 11708A Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AP)

F requency Range (tuned):

Step A ttenuator Flatness (AA)

2.5 MHz to 1.3 GHz

Step A ttenuator Flatness (AP)

Range0 dBm to 127 dBm
Relative P o wer Accuracy 0.5 dB

AM

Rates20 Hz to 100 kHz
Depthto 99%
Accuracy 1% of reading 1 count

F requency Range (carrier):

HP/Agilent 3048A Single Sideband Phase

0.01 to 18 GHz

Sensitivit y:

< 70 dBc at 100 Hz o set

< 78 dBc at 1 kHz o set

< 86 dBc at 10 kHz o set

< 107 dBc at 100 kHz o set

O set F requency Range:

100 Hz to 2 MHz

Amplitude Accuracy:

2 dB to 1 MHz o set

F requency10 MHz
In ternal TimebaseAging
Stability< 1 x 10

10 MHz Standard (A)

F requency RangeDC to 40 GHz

HP/Agilent 33330E External Leveling (P)

Maxim um Input200 mW
P olarit yNegativ e

AM O set (A)

1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Show 25 more labels / short text on this page
Use · Mo del · Amp/Filt Adjustments (A) · Accuracy (A) · Accuracy (AP) · Measuring · HP/Agilent 8902A · Receiver · Phase Noise · Measuremen t · Noise (P) · System · F requency · HP/Agilent 5061A/ · Standard · 10 · /yr · HP/Agilent 5071A · Rate (P) · Crystal · Detector · AM Bandwidth (P) · Amp/Mult Adjustments (A) · 1-50 Required Equipment · Agilent 83640L
Page 82 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 82 · click to enlarge
Page 82

T able 1-8. Required Equipment for 83640L (continued)

Instrument Critical Sp eci cations Recommended

A ttenuator F requency Range10 MHz to 40 GHz

Maxim um Leveled P o wer (P)

Maxim um Input P o wer: 300 mW

A ttenuation20 dB
A ttenuator F requency Range50 MHz to 40 GHz

P o wer Accuracy (P)

Maxim um Input P o wer: 300 mW

A ttenuation20 dB

Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AA)

T o ol Kit No Substitute P/N 08360{60060

In v ertron California

Capacitor 1000 pf P/N 0160-4574 External Leveling (P)

F requency Range2 to 8 GHz

HP/Agilen t 11975A Spurious Signals

Leveled Output P o wer+16 dBm
Mixer F requency Range26.5 to 40 GHz HP/Agilen t 11970A Spurious Signals
Mixer F requency Range20 to 26.5 GHz HP/Agilen t 11970K Spurious Signals

WR42 to APC 3.5 mm (f ) K281C Spurious Signals

WR28 to APC 2.4 mm (f ) R281A Spurious Signals

1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est

P { Manual P erformance T est

Agilent 8360 Required Equipment

Show 25 more labels / short text on this page
Use · Mo del · HP/Agilen t 8490D · Option 020 · HP/Agilen t 8490D · Option 010 · Spurious Signals · (Harmonic) (P) · Amp/Mult Adjustments (A) · Amp/Filt Adjustments (A) · Spurious Signals · Instruments 501TC · (Line Related) (P) · Microwa v e · Ampli er · (Harmonic) (P) · (Harmonic) (P) · (Harmonic) (P) · K{Band Coax · Adapter · (Harmonic) (P) · R{Band Coax · Adapter · (Harmonic) (P) · Agilent 83640L
Page 83 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 83 · click to enlarge
Page 83

The following list of adapters and cables is provided for con v enience. They may b e used in

equipmen t setups for p erformance tests or adjustmen ts.

SMA (m) to SMA (m) adapter 1250-1159

SMA (f ) to SMA (f ) adapter 1250-1158

SMB (m) to SMB (m) adapter 1250-0669

SMB (f ) to SMB (f ) adapter 1250-0672

SMB tee (f ) (m) (m) 1250-1391

3.5 mm (f ) to 3.5 mm (f ) adapter 5061-5311

3.5 mm (f ) to N-t yp e (m) adapter 1250-1745

2.4 mm (f ) to 2.92 mm (f ) adapter 1250-2187

2.4 mm (f ) to 2.4 mm (f ) adapter 1250-2188

2.4 mm (m) to 3.5 mm (f ) adapter 11901C

BNC (f ) to BNC (f ) adapter 1250-0080

BNC (m) to BNC (m) adapter 1250-0216

BNC (f ) to SMA (m) adapter 1250-1200

BNC (f ) to SMB (m) adapter 1250-1237

BNC tee (m) (f ) (f ) 1250-0781

SMB (f ) to BNC (m) exible cable 85680-60093

SMA semi-rigid cable 2 feet 08340-20124

GPIB cable 2 meters HP/Agilent 10833B

BNC cable 8120-2582 · 1-52 Required Equipment · Agilent 83640L

Page 84 of the Agilent Technologies 83640B user manual — ADC (AA)
Page 84 · click to enlarge
Page 84

Agilent 83650L Required Equipment

T able 1-9. Required Equipment for 83650L

Instrument Critical Sp eci cations Recommended

F requency Range10 MHz to 22 GHz

(50 GHz with external mixers)

F requency Span0 Hz, 100 Hz to 22 GHz
Amplitude Range+20 to 100 dBm
Resolution Bandwidth10 Hz to 3 MHz
Video Bandwidth10 Hz to 3 MHz
Log Fidelit y 0.1 dB/dB o v er 0 to

80 dB display (1.0 dB maxim um)

Video OutputDC v oltage prop ortional

to v ertical p osition of trace on display

and API Spurs (A)

Capable of phase lo cking to external

Square W a v e Symmetry (A)

A vs B Sweep Mo de

V ertical Sensitivit y5 mV/Div
Bandwidth100 MHz

AM O set (A)

Controller 4 Mb yte RAM

Step A ttenuator Flatness (AP)

BASIC 5.1

P o wer Flatness and

GPIB

Step A ttenuator Flatness (AA)

ADC (AA)

P o wer Flatness (AA)

P/N 08360-10001 Step A ttenuator Flatness (AP)

(Shipp ed with instrument)

P o wer Flatness and

Step A ttenuator Flatness (AA)

ADC (AA)

P o wer Flatness (AA)

D VM Range 50 to +50 VDC
Accuracy 0.01%

Lo w P o wer SRD Bias (A)

Input Imp edance 10 M

ADC (AA)

1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Agilent 8360 Required Equipment

Show 35 more labels / short text on this page
Use · Mo del · Sp ectrum · HP/Agilen t 8566B · Swept F requency · Analyzer · Accuracy (P) · Spurious Signals · (Harmonic) (P) · Spurious Signals · (Non-harmonic) (P) · Spurious Signals · (Line Related) (P) · F ractional{N Reference · 10 MHz reference · Analog · HP/Agilen t 1740A/ · Amp/Mult Adjustments (A) · Oscilloscop e · HP/Agilen t 54600B · Amp/Filt Adjustments (A) · HP/Agilen t 9836/ · HP/Agilen t 9920/ · HP/Agilen t 310/ · Accuracy (AP) · HP/Agilen t 320 · YO Delay (AA) · Software No Substitute · Accuracy (AP) · YO Delay (AA) · HP/Agilen t 3456A · / · External Leveling (P) · HP/Agilen t 3457A · Agilent 83650L
Page 85 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 85 · click to enlarge
Page 85

T able 1-9. Required Equipment for 83650L (continued)

Instrument Critical Sp eci cations Recommended

In ternal TimebaseAging
BandwidthDC to 300 MHz
Input Imp edance1 M and 50
V ertical Sensitivit y 5 mV/Div
Horizon tal Sensitivit y50 ns/Div

F requency Switching Time (P)

T riggerEv ent T riggerable

10 MHz Standard (A)

Division Ratio10:1 HP/Agilent 10431A Swept F requency

F requency Switching Time (P)

P o wer Meter P o wer Range: 1 W to 100 mW

P o wer Accuracy (P)

Accuracy 0.02 dB

ALC P o wer Level

P o wer Meter P o wer Range: 1 W to 100 mW

HP/Agilent 437B P o wer Flatness (P)

Accuracy 0.02 dB

P o wer Flatness (A)

P o wer Meter P o wer Range: 1 W to 100 mW

HP/Agilent 438A P o wer Flatness and

Accuracy 0.02 dB

Step A ttenuator Flatness (AP)

P o wer Flatness (AA)

Step A ttenuator Flatness (AA)

P o wer Sensor F requency Range: 10 MHz to 2.3 GHz

HP/Agilent 8482A P o wer Flatness (P)

P o wer Range1 W to 100 mW

P o wer Flatness (A)

P o wer Flatness and

Step A ttenuator Flatness (AA)

P o wer Flatness (AA)

Step A ttenuator Flatness (AP)

1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Show 27 more labels / short text on this page
Use · Mo del · Digitizing · Dual Channel · HP/Agilent 54110A/ · Oscilloscop e · HP/Agilent 54111D · Rate (P) · HP/Agilent 54222A/ · Swept F requency · Accuracy (P) · HP/Agilent 54750A/ · HP/Agilent 54752A · Oscilloscop e · Prob es · Accuracy (P) · Amp/Mult Adjustments (A) · Amp/Filt Adjustments (A) · HP/Agilent 436A/ · HP/Agilent 437B/ · Amp/Filt Adjustments (A) · HP/Agilent 438A · Accuracy (A) · Accuracy (AP) · Accuracy (AP) · 1-54 Required Equipment · Agilent 83650L
Page 86 of the Agilent Technologies 83640B user manual — AM
Page 86 · click to enlarge
Page 86

T able 1-9. Required Equipment for 83650L (continued)

Instrument Critical Sp eci cations Recommended

P o wer Sensor F requency Range: 50 MHz to 50 GHz

HP/Agilen t 8487A P o wer Accuracy (P)

P o wer Range1 W to 100 mW

P o wer Flatness (P)

ALC P o wer Level

P o wer Flatness (A)

P o wer Flatness and

Step A ttenuator Flatness (AP)

P o wer Flatness (AA)

Step A ttenuator Flatness (AA)

P o wer Sensor F requency Range: 50 MHz to 50 GHz

HP/Agilen t 8487D Step A ttenuator Flatness (AP)

P o wer Range100 pW to 10 W

Step A ttenuator Flatness (AA)

A ttenuator A ttenuation: 30 dB 0.05 dB at 50 MHz HP/Agilen t 11708A Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AP)

F requency Range (tuned):

Step A ttenuator Flatness (AA)

2.5 MHz to 1.3 GHz

Step A ttenuator Flatness (AP)

Range0 dBm to 127 dBm
Relative P o wer Accuracy 0.5 dB

AM

Rates20 Hz to 100 kHz
Depthto 99%
Accuracy 1% of reading 1 count

F requency Range (carrier):

HP/Agilen t 3048A Single Sideband Phase

0.01 to 18 GHz

Sensitivit y:

< 70 dBc at 100 Hz o set

< 78 dBc at 1 kHz o set

< 86 dBc at 10 kHz o set

< 107 dBc at 100 kHz o set

O set F requency Range:

100 Hz to 2 MHz

Amplitude Accuracy:

2 dB to 1 MHz o set

F requency10 MHz
In ternal TimebaseAging
Stability< 1 x 10

10 MHz Standard (A)

F requency RangeDC to 50 GHz

HP/Agilen t 33330E External Leveling (P)

Maxim um Input200 mW
P olarit yNegativ e

AM O set (A)

1 A { Manual Adjustmen t AA { Automated Adjustmen t AP { Automated P erformance T est

P { Manual P erformance T est

2 Recommended mo del is part of the microw ave test station.

Agilent 8360 Required Equipment

Show 23 more labels / short text on this page
Use · Mo del · Amp/Filt Adjustments (A) · Accuracy (A) · Accuracy (AP) · Measuring · HP/Agilen t 8902A · Receiv er · Phase Noise · Measurement · Noise (P) · System · F requency · HP/Agilen t 5061A/ · Standard · 10 · HP/Agilen t 5071A · Rate (P) · /yr · Crystal · Detector · Amp/Mult Adjustments (A) · Agilent 83650L
Page 87 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 87 · click to enlarge
Page 87

T able 1-9. Required Equipment for 83650L (continued)

Instrument Critical Sp eci cations Recommended

A ttenuator F requency Range10 MHz to 50 GHz

Maxim um Leveled P o wer (P)

Maxim um Input P o wer: 300 mW

Pulse P erformance (P)

A ttenuation20 dB
A ttenuator F requency Range50 MHz to 50 GHz

P o wer Accuracy (P)

Maxim um Input P o wer: 300 mW

A ttenuation20 dB

Step A ttenuator Flatness (AA)

Step A ttenuator Flatness (AA)

T o ol Kit No Substitute P/N 08360{60060

In v ertron California

Capacitor 1000 pf P/N 0160-4574 External Leveling (P)

Mixer F requency Range26.5 to 40 GHz HP/Agilent 11970A Spurious Signals
F requency Range2 to 8 GHz

HP/Agilent 11975A Spurious Signals

Leveled Output P o wer+16 dBm
Mixer F requency Range20 to 26.5 GHz HP/Agilent 11970K Spurious Signals
Mixer F requency Range40 to 50 GHz HP/Agilent 11970Q Spurious Signals

WR42 to APC 3.5 mm (f ) K281C Spurious Signals

WR28 to APC 2.4 mm (f ) R281A Spurious Signals

WR22 to APC 2.4 mm (f ) Q281A Spurious Signals

1 A { Manual Adjustmen t AA { Automated Adjustment AP { Automated P erformance T est

P { Manual P erformance T est

Show 30 more labels / short text on this page
Use · Mo del · HP/Agilent 8490D · Option 020 · HP/Agilent 8490D · Option 010 · Spurious Signals · (Harmonic) (P) · Amp/Mult Adjustments (A) · Amp/Filt Adjustments (A) · Spurious Signals · Instruments 501TC · (Line Related) (P) · (Harmonic) (P) · Microwa v e · Ampli er · (Harmonic) (P) · (Harmonic) (P) · (Harmonic) (P) · K{Band Coax · Adapter · (Harmonic) (P) · R{Band Coax · Adapter · (Harmonic) (P) · Q{Band Coax · Adapter · (Harmonic) (P) · 1-56 Required Equipment · Agilent 83650L
Page 88 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 88 · click to enlarge
Page 88

The follo wing list of adapters and cables is provided for con v enience. They may b e used in

equipmen t setups for p erformance tests or adjustmen ts.

SMA (m) to SMA (m) adapter 1250-1159

SMA (f ) to SMA (f ) adapter 1250-1158

SMB (m) to SMB (m) adapter 1250-0669

SMB (f ) to SMB (f ) adapter 1250-0672

SMB tee (f ) (m) (m) 1250-1391

3.5 mm (f ) to 3.5 mm (f ) adapter 5061-5311

3.5 mm (f ) to N-t yp e (m) adapter 1250-1745

2.4 mm (f ) to 2.92 mm (f ) adapter 1250-2187

2.4 mm (f ) to 2.4 mm (f ) adapter 1250-2188

2.4 mm (m) to 3.5 mm (f ) adapter 11901C

BNC (f ) to BNC (f ) adapter 1250-0080

BNC (m) to BNC (m) adapter 1250-0216

BNC (f ) to SMA (m) adapter 1250-1200

BNC (f ) to SMB (m) adapter 1250-1237

BNC tee (m) (f ) (f ) 1250-0781

SMB (f ) to BNC (m) exible cable 85680-60093

SMA semi-rigid cable 2 feet 08340-20124

GPIB cable 2 meters HP/Agilent 10833B

Agilent 8360 Required Equipment

BNC cable 8120-2582 · Agilent 83650L

Page 89 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 89 · click to enlarge
Page 89
Page 89 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 90 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNN
Page 90 · click to enlarge
Page 90

P erformance T ests

How to Use This Chapter

Use the pro cedures in this c hapter to test the electrical p erformance of the syn thesizer. These

tests do not require access to the in terior of the instrumen t.

The syn thesizer m ust w arm up for at least one hour b efore the electrical sp eci cations are

v alid.

F old-out men u maps are lo cated in Chapter 6. Some men us ha v e more than one page of

softkeys. Select the

NNNNNNNNNNNNNN

more softkey to view the next page of softkeys.

NNNNNNNNNNNNNN

more is not included in

the k eystrokes given in these pro cedures.

In all cases where y ou are instructed to preset the syn thesizer, use the factory

preset mo de only .

F or assurance that most of the in ternal functions of the instrumen t w ork (without testing for

sp eci cations), see \Op erator's Chec k," in y our syn thesizer's User's guide.

Agilent 8360 P erformance T ests 2-1

Menu Maps · Operator's Check

Page 91 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 91 · click to enlarge
Page 91

List of T ests

The following is a list of this c hapter's tests and the page n um b er that eac h test is lo cated on:

Op eration V eri cation T ests : 2-5

P erformance T ests : 2-8

1. In ternal Timebase: Aging Rate : 2-9
2. Sw ept F requency Accuracy : 2-13
3. F requency Switching Time : 2-23
4. P ow er Accuracy : 2-30
5. P ow er Flatness : 2-33
6. Maximum Lev eled P ow er : 2-36
7. External Lev eling : 2-39
8. Spurious Signals (Harmonics & Subharmonics) : 2-42
9. Spurious Signals (Non-Harmonic) : 2-53
10. Spurious Signals (Line-Related) : 2-59
11. Single Sideband Phase Noise : 2-63
12. Pulse Mo dulation On/O Ratio y : 2-68
13. Pulse P erformance y : 2-71
14. Pulse P erformance (Alternate Pro cedure) y : 2-79
15. Pulse Mo dulation Video F eedthrough y : 2-90
16. AM Accuracy y : 2-99
17. AM Bandwidth y : 2-107
18. AM Dynamic Range y : 2-111
19. FM Accuracy y : 2-118
20. FM Bandwidth y : 2-122
21. Maxim um FM Deviation y : 2-128
22. In ternal Pulse Accuracy y (Option 002) : 2-135
23. Mo dulation Meter y (Option 002) : 2-137

P erformance T ests for B-Series Only

The p erformance tests that apply to the

Agilent 8360 B-series swept signal generators

only are marked with the following symb ol: y

Operation V erification T ests

T o meet the needs of most incoming insp ections

(80% v eri cation) , and provide reasonable assurance that the instrumen t w orks, see the

\Op eration V eri cation" section of this c hapter.

Automated P erformance T ests

The automated p erformance tests av ailable at time-of-shipment are included in Chapter 5.

The Step A tten uator test is lo cated in Chapter 5.

2-2 P erformance T ests Agilent 8360

Page 92 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 92 · click to enlarge
Page 92

The equipmen t required to p erform the tests in this c hapter is listed in Chapter 1. Y ou may

use an y equipmen t that meets the critical sp eci cations given.

T est records are supplied at the end of this c hapter. Use a test record when y ou p erform a full

calibration of y our syn thesizer. This form provides a tabulated index of the p erformance tests,

their acceptable limits, the measuremen t uncertain ty of eac h test, and a column to record

actual measuremen ts.

There are sev eral test records. Be sure y ou use the one designated (at the top) for y our

syn thesizer. Eac h test record is sho wn b elow. The page n um b er of eac h test record is also

included for easy access.

T est Record for Agilent 83620B /22B : 2-139

T est Record for Agilent 83623B /24B : 2-147

T est Record for Agilent 83630B : 2-155

T est Record for Agilent 83640B : 2-163

T est Record for Agilent 83650B : 2-171

T est Record for Agilent 83623L : 2-179

T est Record for Agilent 83630L : 2-184

T est Record for Agilent 83640L : 2-189

T est Record for Agilent 83650L : 2-194

P erform the tests in the order that they app ear.

P erform the tests in this c hapter at least once ev ery 24 mon ths.

Agilent 8360 P erformance T ests 2-3

Show 4 more labels / short text on this page
Equipment Required · T est Record · T est Sequence · Calibration Cycle
Page 93 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 93 · click to enlarge
Page 93

The following pro cedures meet the needs of most incoming insp ections (80% v eri cation), and

provide a reasonable assurance that the instrumen t is functioning prop erly . The Op eration

V eri cation F orm is lo cated b elow.

P erform these op eration v eri cation pro cedures in the order given and record the pass/fail

results in the Op eration V eri cation F orm.

T able 2-1. Operation V erification Form

T est F acilit y Date

Serial Numb er T ested By

Options Firmw are Revision

T est Results (P ass/F ail)

1. Self-T ests
2. P ow er Accuracy
3. P ow er Flatness
4. Maxim um Lev eled P ow er

2-4 P erformance T ests

Operation V erification · Mo del Customer · Operation V erification

Page 94 of the Agilent Technologies 83640B user manual — ALL SELF-TESTS HAVE PASSED!
Page 94 · click to enlarge
Page 94
1. Self-tests
1. P erform the full self-tests con tained in the instrumen t rm w are. No additional test

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

equipmen t is required. Press

SERVICE

Selftest (Full) .

The syn thesizer executes a series of self-tests. When completed, the follo wing message is

display ed if all the tests passed:

ALL SELF-TESTS HAVE PASSED!

Occasionally , the self-tests may fail when it is run remotely or when external

cables are connected to the instrumen t. If the self-tests fail, remov e all cables

and run the self-tests from the front panel. If the self-tests pass under these

conditions, the instrumen t is op erating prop erly .

External connections to the instrumen t during self-test may not allo w the lines

to c hange conditions within the limits of the self-test's parameters.

2. Record Pass on the op eration v eri cation form and con tinue to the next test, \Pow er

Accuracy ."

If the syn thesizer fails one or more self-tests, a di eren t message is display ed listing the

most indep enden t test that failed and a reference to the b est en try p oint in to the A gilent

T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series Swept CW Gener ator

T r oublesho oting Guide . Record F ail on the op eration v eri cation form and see the A gilent

T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series Swept CW Gener ator

T r oublesho oting Guide . After repair, rep eat op eration v eri cation from the b eginning.

2. P o w er Accuracy
1. F ollo w the pro cedure for this p erformance test given in this c hapter.
2. If the syn thesizer passes this test, record Pass on the op eration v eri cation form and

con tinue to the next test, \Pow er Flatness."

If the syn thesizer fails, record F ail on the op eration v eri cation form and then follo w the

\In Case of Diculty" instructions given for this p erformance test. After repair, rep eat

op eration v eri cation from the b eginning.

Agilent 8360 P erformance T ests

· Operation V erification

Page 95 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 95 · click to enlarge
Page 95
3. P o w er Flatness
1. F ollo w the pro cedure for this p erformance test given in this c hapter.
2. If the syn thesizer passes this test, record Pass on the op eration v eri cation form and

con tinue to the next test, \Maxim um Lev eled P ow er."

If the syn thesizer fails, record F ail on the op eration v eri cation form and then follow the

\In Case of Diculty" instructions given for this p erformance test. After repair, rep eat

op eration v eri cation from the b eginning.

4. Maximum Lev eled P o w er
1. F ollo w the pro cedure for this p erformance test given in this c hapter.
2. If the syn thesizer passes this test, record Pass on the op eration v eri cation form.

Op eration v eri cation is completed.

If the syn thesizer fails, record F ail on the op eration v eri cation form and then follow the

\In Case of Diculty" instructions given for this p erformance test. After repair, rep eat

op eration v eri cation from the b eginning.

2-6 P erformance T ests

Operation V erification

Page 96 of the Agilent Technologies 83640B user manual — PERF ORMANCE TESTS
Page 96 · click to enlarge
Page 96

PERF ORMANCE TESTS

1. Internal Timebase: Aging Rate

This pro cedure c hec ks the accuracy of the in ternal timebase. The time required for a sp eci c

phase c hange is measured b oth b efore and after a sp eci ed w aiting p erio d. The aging rate is

in v ersely prop ortional to the absolute v alue of the di erence in the measured times.

The ov erall accuracy of the in ternal timebase is a function of:

TBC AR TE LE where:

TBC = timebase calibration

TE = temperature effects

AR = aging rate

LE = line effects

After the timebase is adjusted, the timebase frequency should stay within the aging rate if the

follo wing things happ en:

The timebase ov en do es not co ol do wn.

The instrumen t k eeps the same orientation with resp ect to the earth's magnetic eld.

The instrumen t stays at the same altitude.

The instrumen t do es not receiv e an y mec hanical sho c k.

If the timebase ov en co ols (the instrumen t is disconnected from A C p ow er), y ou may ha v e to

readjust the timebase frequency after a new w armup cycle. T ypically , ho w ev er, the timebase

frequency returns to within 1 Hz of the original frequency .

The in ternal timebase can b e tested after reconnecting A C p ow er for 10

minutes, but for b est accuracy , test again after the instrumen t has b een on or

in standb y condition for 24 hours.

Agilent 8360 P erformance T ests

Internal TimebaseAging Rate

Description and Procedure

Page 97 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 97 · click to enlarge
Page 97

F requency c hanges due either to a c hange in orientation with resp ect to the earth's magnetic

eld, or to a c hange in altitude, usually go aw a y when the instrumen t is returned to its

original p osition. A frequency c hange due to mec hanical sho c k usually app ears as a xed

frequency error.

1. Connect the equipmen t as sho wn in Figure 2-1 . Preset all instrumen ts and let them w arm

up for at least one hour.

If the oscilloscop e do es not ha v e a 50 input imp edance, connect c hannel 1

through a 50 feedthrough.

Agilent Technologies 83640B — 08360125.pdf diagram (page 97)
Fig. 4 · 08360125.pdf
Figure 2-1. Internal TimebaseAging Rate T est Setup

Digitizing Oscilloscop e HP/Agilent 54111D

F requency Standard HP/Agilent 5061A

2-8 P erformance T ests

Internal TimebaseAging Rate

Equipment

Page 98 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 98 · click to enlarge
Page 98
2. On the oscilloscop e, adjust the external triggering for a display of the 10MHz REF

OUTPUT signal from the syn thesizer:

Channel 1:

Channel 2:

Timebase:

Delay Reference At center

Trigger:

Trig 3 Level 100 mV

Trigger Src Trig 3

Display:

Display Mode Real time

3. Monitor the time and the display . Note the time required for a 360
phase c hangeT1 =
4. W ait 3 to 24 hours. Note ho w long y ou w aited: T2 = (h)
5. Repeat step 3. Record the phase c hange time: T3 = (s)

Agilent 8360 P erformance T ests

Internal TimebaseAging Rate
Show 13 more labels / short text on this page
Display On · Volts/Division 120 mV · Input Coupling dc · Input Impedance 50 · Display Off · Time/Division 5 ns · Delay 0 s · Sweep Trig'd · Trigger Mode Edge · Input Impedance 50 · Input Coupling dc · · (s)
Page 99 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 99 · click to enlarge
Page 99
6. Calculate the aging rate as follo ws:

Aging Rate = (1 cycle/10 MHz) (1/T1 1/T3) (24 hours/T2)

ExampleT1 = 351 seconds

T2 = 3 hours

T3 = 349 seconds

= (1 cycle/10 MHz) (1/351s 1/349s) (24h/3h)

= 1.306x10

7. En ter the aging rate on the test record.

If the absolute frequency of the standard and of the timebase oscillator are

extremely close, y ou can reduce the measuremen t time (T1 and T3) b y

measuring the time required for a phase c hange of less than 360

. In step 6,

c hange 1 cycle to 0.5 cycle for 180

, or 0.25 cycle for 90

In Case of Difficulty

1. Ensure that the instrumen ts ha v e w armed up long enough and that en vironmental

conditions ha v e not c hanged throughout the test.

2. If the frequency standard and the in ternal standard are v ery di eren t in frequency , the time

required for a 360

phase shift is to o short for an accurate measuremen t. If the 360

shift takes less than tw o minutes, p erform the \10 MHz Standard" adjustmen t.

3. Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

2-10 P erformance T ests

Internal TimebaseAging Rate
Show 10 more labels / short text on this page
11 · per day · · · · Related Adjustments · 10 MHz Standard · · · phase
Page 100 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 100 · click to enlarge
Page 100
2. Sw ept Frequency Accuracy

With the syn thesizer in swept mo de, the sp ectrum analyzer is set to zero span at the

measuremen t frequency . As the syn thesizer sweeps through the sp ectrum analyzer frequency

setting, a signal is generated on the sp ectrum analyzer's video output that is input to the

oscilloscop e.

The syn thesizer's TRIGGER OUTPUT, used to trigger the oscilloscop e, is a series of 1601

pulses, ev enly spaced during the sweep. The oscilloscop e is triggered on the pulse that

represen ts the desired measuremen t frequency , and the sp ectrum analyzer is tuned to display

the video output on the oscilloscop e.

1. Connect the equipmen t as sho wn in Figure 2-2. Preset all instrumen ts and let them w arm

up for at least one hour.

Agilent Technologies 83640B — 08360125.pdf diagram (page 100)
Fig. 5 · 08360125.pdf

Figure 2-2. Sw ept Frequency Accuracy T est Setup

Sp ectrum Analyzer HP/Agilent 8566B

Digitizing Oscilloscop e HP/Agilent 54111D

Oscilloscop e Prob es HP/Agilent 10431A

2. On the sp ectrum analyzer, set the cen ter frequency to the rst cen ter frequency listed in

T able 2-2. Then set:

Agilent 8360 P erformance T ests

Sw ept Frequency Accuracy

Description and Procedure · Equipment

Page 101 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 101 · click to enlarge
Page 101
Span0 Hz
Reference Level0 dBm
Scale Log10 dB/div
Resolution Bandwidth100 kHz
3. Set the oscilloscop e as follows:

T rigger 3 is a trigger enable that ensures that c hannel 2 (the true trigger)

triggers only on a forward sweep after the sp eci ed n um b er of ev en ts.

Channel 1:

Input Impedance 1 M

Channel 2:

Input Impedance 1 M

2-12 P erformance T ests

Sw ept Frequency Accuracy

Show 8 more labels / short text on this page
Display On · Volts/Division 300 mV · Input Coupling dc · Offset 1 V · Display On · Volts/Division 1 V · Offset 2 V · Input Coupling dc
Page 102 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 102 · click to enlarge
Page 102

Timebase:

Time/Division 25 s

Delay Reference At center

Trigger:

Trigger Src Chan 2

Trigger Level 1.6 V

Trigger Src Trig 3

Trigger Level 1.6 V

Trigger After Positive Edge

Trigger On Trig 3

Trigger On 33 events

Trigger Of Positive edge

Trigger On Channel 2

Display:

Display Mode Real time

4. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USER CAL

Swp Span Cal Always (asterisk on)

ST ART

First start frequency in T able 2-2

First stop frequency in T able 2-2

STOP

Set the frequency range rst, or y ou cannot set the sweep time prop erly .

CONT

SWEEP TIME

(F or Option 006, set to 150 ms)

POWER LEVEL

F or the 83640B /L and 83650B /L only , also press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SERVICE

Disable Doubler (asterisk on)

Agilent 8360 P erformance T ests

Sw ept Frequency Accuracy

Show 22 more labels / short text on this page
Delay 0 s · Sweep Triggered · Trigger Mode Edge · Trigger Mode Events · · · Freq Cal Menu · · · · · · 100 · ms · · · 0 · dBm · · · · Tools Menu
Page 103 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 103 · click to enlarge
Page 103

T able 2-2. Sw ept Frequency Accuracy Instrument Settings

Agilent Technologies 83640B — 08360125.pdf diagram (page 103)
Fig. 6 · 08360125.pdf
5. T une the sp ectrum analyzer cen ter frequency to nd and cen ter the video signal on the

oscilloscop e. (See Figure 2-3 .) Align the video input with the trigger output at cen ter

screen.

2-14 P erformance T ests

Sw ept Frequency Accuracy

Page 104 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 104 · click to enlarge
Page 104
Agilent Technologies 83640B — 08360125.pdf diagram (page 104)
Fig. 7 · 08360125.pdf

Figure 2-3. Video Signal on the Oscilloscope

6. Note the nal cen ter frequency setting required to cen ter the video signal.
7. Record the di erence b et w een the initial cen ter frequency setting and the v alue noted in

step 6 in T able 2-2 as frequency error.

8. Repeat steps 5 through 7 for the remaining instrumen t settings in T able 2-2 and T able 2-3

through T able 2-6 as appropriate for eac h instrumen t mo del.

9. Calculate the frequency error as a p ercen t of the sweep width as follows for eac h of the

results in T able 2-2 and T able 2-3 through T able 2-6 :

x100 = p ercent

stop frequency start frequency

10. Record the w orst case v alue on the test record.

F or 83620B /22B/24B and 83623B /L only:

T able 2-3. Additional Instrument Settings

T rigger Ev ents

2 20 2.10125 10

2 20 6.8825 435

2 20 7.1525 459

2 20 13.25 1001

2 20 13.79 1049

2 20 19.775 1581

Agilent 8360 P erformance T ests

Sw ept Frequency Accuracy

Show 10 more labels / short text on this page
frequency error · Synthesizer · Sp ectrum Analyzer · Oscilloscope · F requency · F requency (GHz) · Center F requency · Error Percent · (GHz) · Start Stop
Page 105 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 105 · click to enlarge
Page 105

F or 83630B /L only:

T able 2-4. Additional Instrument Settings

Agilent Technologies 83640B — 08360125.pdf diagram (page 105)
Fig. 8 · 08360125.pdf

F or 83640B /L only:

T able 2-5. Additional Instrument Settings

Agilent Technologies 83640B — 08360125.pdf diagram (page 105)
Fig. 9 · 08360125.pdf

2-16 P erformance T ests

Sw ept Frequency Accuracy

Page 106 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 106 · click to enlarge
Page 106

F or 83650B /L only:

T able 2-6. Additional Instrument Settings

Agilent Technologies 83640B — 08360125.pdf diagram (page 106)
Fig. 10 · 08360125.pdf

YO Driver +10 V Reference

YO Driver Gain and Linearity

YO Delay Comp ensation

Sw eep Ramp Calibration

In Case of Difficulty

1. V erify that the sp ectrum analyzer frequency is accurate. If necessary , calibrate the

frequency with the syn thesizer's 10 MHz reference connected to the sp ectrum analyzer's

external reference.

2. Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

Agilent 8360 P erformance T ests

Sw ept Frequency Accuracy

Related Adjustments

Page 107 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 107 · click to enlarge
Page 107
3. Frequency Switching Time

The syn thesizer's Z-AXIS BLANK/MARKER output go es active high when a c hange in

frequency is initiated, and returns low when the syn thesizer settles at the new frequency .

Using an oscilloscop e to monitor this output, the frequency switching time is measured in

CW, step sweep, and frequency list mo des.

1. Connect the equipmen t as sho wn in Figure 2-4 . Preset b oth instrumen ts and let them

w arm up for at least one hour.

Agilent Technologies 83640B — 08360125.pdf diagram (page 107)
Fig. 11 · 08360125.pdf

Figure 2-4. Frequency Switching Time T est Setup

Digitizing Oscilloscop e HP/Agilent 54111D

Oscilloscop e Prob es HP/Agilent 10431A

2-18 P erformance T ests

Description and Procedure · Equipment · Frequency Switching Time

Page 108 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 108 · click to enlarge
Page 108

CW Frequency Switching Time (Across Bandswitch P oints)

2. On the oscilloscop e, set:

Channel 1:

Input Impedance 1 M

Channel 2:

Timebase:

Delay Reference At left

Delay 10 ms

Trigger:

Trigger Src Chan 1

Trigger Level 1.6 V

Display:

3. On the syn thesizer, set the rst Initial CW F r e quency in T able 2-7 .
4. On the oscilloscop e, clear the display . The oscilloscop e should display Awaiting Trigger .
5. On the syn thesizer, set the rst Se c ond CW F r e quency in T able 2-7 .

T able 2-7. CW Frequency Switching Time Settings

Initial CW F requency (GHz) Second CW F requency (GHz) Pulse Width

13.5 50

1 Set to 2 GHz (low est start frequency) for synthesizers not capable of 0.01 GHz.

2 Set to the highest stop frequency for synthesizers not capable of 50 GHz.

6. On the oscilloscop e, note the pulse width of the Z-axis blank/markers signal. Record this

v alue as pulse width in T able 2-7 .

7. Clear the oscilloscop e display and rep eat steps 3 through 6 for the remaining frequencies

in T able 2-7.

8. On the test record, record the maxim um pulse width from T able 2-7 .

Agilent 8360 P erformance T ests

Show 11 more labels / short text on this page
Display On · Preset TTL · Input Coupling dc · Display Off · Time/Division 10 ms · Sweep Triggered · Trigger Mode Edge · Trigger Slope Pos · Display Mode Repetitive · Averaging Off · Frequency Switching Time
Page 109 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 109 · click to enlarge
Page 109

Stepped Sw eep Frequency Switching Time (Within a Frequency Band)

9. On the syn thesizer, press:

FREQUENCY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

2.4

ST ART

STOP

6.9

SINGLE

SWEEP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MENU

SINGLE

10. On the oscilloscop e, set:

Channel 1:

Timebase:

Delay Reference At left

Delay 2 ms

Trigger:

Trigger After Negative edge

Trigger Src Chan 1

Trigger On 1 events

Trigger Of Positive edge

Trigger On Channel 1

Display:

11. Set the oscilloscop e for a single sweep at the next trigger. Press

. The

oscilloscop e should display Awaiting Trigger.

12. On the syn thesizer, initiate a stepp ed sweep. Press

SINGLE

.

13. On the oscilloscop e, measure the p ositiv e pulse width and note the v alue .
14. Incremen t the trigger to 20 ev en ts and rep eat steps 11 through 13.

Pulse width v alue .

15. Incremen t the trigger to 45 ev en ts and rep eat steps 11 through 13.

Pulse width v alue .

16. Record the w orst case v alue from steps 13 through 15 on the test record.

Frequency List Frequency Switching Time (Across Bandswitch P oints)

17. On the syn thesizer, delete an y en tries in the frequency list men u. Press:

FREQUENCY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNN

MENU

On the syn thesizer, press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Enter List Freq and en ter the following frequencies in the

frequency list men u:

2-20 P erformance T ests

Show 37 more labels / short text on this page
· Step Swp Menu · Step Size · 100 · · MHz · Step Dwell · 10 · · ms · · · · GHz · · · GHz · · · · · Sweep Mode Step · · · Display On · Preset TTL · Input Coupling dc · Time/Division 2 ms · Sweep Triggered · Trigger Mode Events · Display Mode Repetitive · Stop/Single · · List Menu · Delete Menu · All · Frequency Switching Time
Page 110 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 110 · click to enlarge
Page 110

F requency List for Syn thesizers with 0.01 GHz Capability

F requency (GHz):

Disregard if b ey ond y our syn thesizer's capabilit y .

F requency List for Syn thesizers without 0.01 GHz Capability

F requency (GHz):

Disregard if b ey ond y our syn thesizer's capabilit y .

19. On the oscilloscop e, set:

Channel 1:

Channel 2:

Timebase:

Delay Reference At left

Delay 2 ms

Trigger:

Trigger After Negative edge

Trigger Src Channel 1

Trigger On 1 events

Trigger Of Positive edge

Trigger On Channel 1

Display:

Agilent 8360 P erformance T ests

Show 9 more labels / short text on this page
Display On · Preset TTL · Input Coupling dc · Display Off · Time/Division 5 ms · Sweep Triggered · Trigger Mode Events · Display Mode Repetitive · Frequency Switching Time
Page 111 of the Agilent Technologies 83640B user manual — SINGLE
Page 111 · click to enlarge
Page 111
20. On the syn thesizer, press:

SINGLE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SWEEP

MENU

SINGLE

21. On the oscilloscop e, set a single sweep at the next trigger. Press

. The

oscilloscop e should display Awaiting Trigger.

22. On the syn thesizer, initiate a frequency list sweep. Press

.

SINGLE

23. On the oscilloscop e, measure the p ositiv e pulse width and note the v alue:

Pulse Width V alue

1 ev en t

2 ev en ts

3 ev en ts

4 ev en ts

5 ev en ts

6 ev en ts

1 Not applicable for all mo dels. (Each even t corresp onds to a

change in frequency in the frequency list.)

24. Incremen t the trigger on ev en ts and rep eat steps 21 and 22 for a trigger on ev en ts up to

and including 6.

25. Record the w orst case v alue from step 23 on the test record.

F ractional-N V CO

YO Lo op Gain

In Case of Difficulty

1. V erify that the oscilloscop e triggers when stepping b et w een tw o frequencies. If necessary ,

adjust the trigger lev el on the oscilloscop e.

2. Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

2-22 P erformance T ests

Show 10 more labels / short text on this page
· · · Sweep Mode List · · · Stop/Single · Related Adjustments · Sampler Assem bly · Frequency Switching Time
Page 112 of the Agilent Technologies 83640B user manual — POWER LEVEL
Page 112 · click to enlarge
Page 112
4. P o w er Accuracy

Using a p ow er meter, c hec k the p ow er accuracy of the syn thesizer at sev eral CW frequencies.

A t eac h frequency , v erify that the actual output p ow er is within sp eci cation ov er the full

dynamic range of the ALC lo op.

1. T urn on the equipmen t sho wn in Figure 2-5 and let them w arm up for at least one hour.

PRESET

on the syn thesizer.

2. On the p ow er meter:

a. Zero and calibrate the p ow er meter/sensor.

b. Set to dBm mo de.

3. Connect the equipmen t as sho wn in Figure 2-5.
4. T o ac hiev e p eak p ow er, turn on RF p eaking. Press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USER CAL

Peak RF Always (asterisk on).

Agilent Technologies 83640B — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 112)
Fig. 12

Figure 2-5. P o w er Accuracy T est Setup

P ow er Meter HP/Agilent 436A/7A/8A

P ow er Sensor HP/Agilent 8487A (83640B /L and 83650B/L)

P ow er Sensor HP/Agilent 8485A (83620B /22B/24B, 83623B/L,

A tten uator 10 dB HP/Agilent 8490D Option 010 (83640B/L and 83650B/L)

A tten uator 10 dB HP/Agilent 8493C Option 010 (83620B/22B/24B,

5. On the syn thesizer, press:

CW

En ter the rst CW v alue in T able 2-8 .

PO WER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Uncoupl Atten (Option 001 only)

POWER LEVEL

RF ON/OFF

O (amb er light o )

6. Zero the p ow er meter.
7. On the syn thesizer, turn the RF on (amb er light on).

Agilent 8360 P erformance T ests

P o w er Accuracy

Show 21 more labels / short text on this page
Description and Procedure · Press · · · Tracking Menu · Equipment · 83623B/L, and 83630B/L) · 83623B/L, and 83630B/L) · · · Up/Dn Power · 1 · · dB · · · 10 · · dBm · ·
Page 113 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 113 · click to enlarge
Page 113

T able 2-8. P o w er Accuracy Frequencies

at 10 dBm

at > 10 dBm

1 Disregard if b eyond your synthesizer's capability .

8. On the p ow er meter, set the p ow er sensor calibration factor for the frequency to b e

measured.

9. Note the di erence b et w een the p ow er meter reading and the p ow er v alue set on the

syn thesizer. W rite this v alue do wn on a separate piece of pap er.

F or p ow er lev els of 10 dBm and ab ov e, measure the p ow er at 10 dBm, then set

the attenuator to the 10 dB p osition (use a 10 dB attenuator for syn thesizers

without a step attenuator.) Record the di erence b et w een the measuremen ts:

(approximately 10 dB). Then, with the attenuator still set to

10 dB, complete the remaining measuremen ts. Eac h of these measuremen ts

should b e o set b y the di erence in dB that y ou measured.

10. On the syn thesizer, use the up

* k ey to incremen t the p ow er lev el 1 dB.

11. Repeat steps 8 through 10, to the maximum sp eci ed p ow er lev el of y our syn thesizer.

Record the w orst case measured di erence (greatest duration) for this frequency in

T able 2-8 for p ow er lev els 10 dBm and, if y our syn thesizer is capable, for p ow er lev els >

10 dBm.

12. On the syn thesizer, set the p ow er lev el to 10 dBm.
13. Repeat steps 8 through 11 for the remaining frequencies in T able 2-8 not to exc e e d the

maximum sp eci ed frequency of y our syn thesizer. Remem b er to delete the 10 dB of

attenuation for p ow er lev els b elow 0 dBm.

14. Record the w orst case measured v alues for eac h frequency range and p ow er range on the

test record.

Mo dulator O set and Gain

ALC P ow er Lev el Accuracy

ALC P ow er Flatness

2-24 P erformance T ests

P o w er Accuracy

Show 5 more labels / short text on this page
F requency · Measured Di erence · Measured Di erence · (GHz) · Related Adjustments
Page 114 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 114 · click to enlarge
Page 114

In Case of Difficulty

Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

Agilent 8360 P erformance T ests

P o w er Accuracy

Page 115 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 115 · click to enlarge
Page 115
5. P o w er Flatness

This pro cedure uses the user atness correction arra y to automatically measure p ow er atness.

The p ow er meter is connected directly to the syn thesizer's RF output. The syn thesizer

con trols the p ow er meter via GPIB while the p ow er meter measures the RF output. (There

cannot b e another con troller on the GPIB during this test.) If the syn thesizer has a step

attenuator, it is set to 0 dB so that an y input in to the atness array indicates the RF output

p ow er atness.

This p erformance test requires an HP/Agilen t 437B p ow er meter. The correct

p ow er sensor calibration factors m ust b e loaded and selected.

This pro cedure deletes an y existing user atness correction arra y .

1. T urn on the equipmen t sho wn in Figure 2-6. Connect the p ow er sensor for the frequency

range b eing calibrated. Do not connect the p ow er sensor to the syn thesizer y et. Preset all

instrumen ts and let them w arm up for at least one hour.

Agilent Technologies 83640B — 08360125.pdf diagram (page 115)
Fig. 13 · 08360125.pdf

Figure 2-6. P o w er Flatness T est Setup

P ow er Meter HP/Agilent 437A

P ow er Sensor HP/Agilent 8482A

P ow er Sensor HP/Agilent 8487A (83640B/L and 83650B/L)

P ow er Sensor HP/Agilent 8485A (83620B/22B/24B, 83623B/L,

Lo w Band P o w er Flatness

Disr e gar d if b eyond your synthesizer's c ap ability.

2. In the syn thesizer's atness men u:

a. Delete all en tries in the user atness correction array . Press:

PO WER

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

b. Set the auto ll stop frequency to 2 GHz:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Auto Fill Stop

2-26 P erformance T ests

P o w er Flatness

Show 10 more labels / short text on this page
Description and Procedure · Equipment · and 83630B/L) · · Fltness Menu · Delete Menu · Delete All · 2 · GHz ·
Page 116 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 116 · click to enlarge
Page 116

c. Set the auto ll start frequency to 15 MHz:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Auto Fill Start

PRIOR

d. Set the auto ll incremen t to 10 MHz:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Auto Fill Incr

3. On the p ow er meter, load the calibration factors for the p ow er sensor.

a. Zero and calibrate the p ow er meter/sensor.

b. Connect the p ow er sensor to the syn thesizer's RF OUTPUT as sho wn in Figure 2-6 .

4. Set the syn thesizer to measure the p ow er correction v alues and generate a new user atness

array for the frequency range en tered. Press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

When the atness correction array is completed, the men u is returned.

5. Scroll through the user atness correction v alues using the up/do wn arrow k eys. Note

the maximum and minimum correction v alues. The di erence b et w een them is the p ow er

atness measuremen t. Record the v alue on the test record and compare the v alue to the

sp eci cation.

6. Delete the atness correction arra y . Press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PO WER

MENU

High Band P o w er Flatness

7. With the appropriate p ow er sensor for this frequency range, rep eat steps 2 through 6 for:

A 20 GHz stop frequency .

A 2.05 GHz start frequency .

A 100 MHz incremen t.

Millimeter Band P o w er Flatness > 20 GHz and 40 GHz

Disr e gar d if b eyond your synthesizer's c ap ability.

8. With the appropriate p ow er sensor for this frequency range, rep eat steps 2 through 6 for:

A 40 GHz stop frequency (for 83630B /L only , set to 26.5 GHz).

A 20.05 GHz start frequency .

A 100 MHz incremen t.

Millimeter Band P o w er Flatness > 40 GHz

Disr e gar d if b eyond your synthesizer's c ap ability.

9. With the appropriate p ow er sensor for this frequency range, rep eat steps 2 through 6 for:

A 50 GHz stop frequency .

A 40.05 GHz start frequency .

A 100 MHz incremen t.

P ow er Flatness

In Case of Difficulty

1. The p ow er meter m ust b e an HP/Agilent 437B.
2. The correct calibration factors for the p ow er sensor m ust b e loaded.
3. Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

Agilent 8360 P erformance T ests

P o w er Flatness

Show 15 more labels / short text on this page
· · 15 · · MHz · 10 · MHz · · Mtr Meas Menu · Measure Corr All · · Fltness Menu · Delete Menu · Delete All · Related Adjustments
Page 117 of the Agilent Technologies 83640B user manual — RF ON/OFF
Page 117 · click to enlarge
Page 117
6. Maximum Lev eled P o w er

This p erformance test is not v alid unless the syn thesizer meets b oth its p ow er

accuracy and p ow er atness sp eci cations. P erform those tests rst.

The unlev eled status indicator is display ed when the instrumen t is unlev eled as the syn thesizer

sweeps ov er sp eci c frequency ranges in fast con tinuous sweep, and fast and slow single sweep

op eration. Because of the syn thesizer's p ow er accuracy and atness p erformance, a p ow er

meter is not required for this measuremen t. The following pro cedure tests the most likely

w orst case situations for maximum lev eled p ow er.

1. Set up and turn on the equipmen t sho wn in Figure 2-7. Preset the syn thesizer and let it

w arm up for at least one hour.

The 20 dB attenuator provides a go o d match on the RF output. If the

syn thesizer has a step attenuator, y ou can simulate a go o d match b y

decoupling the step attenuator from the ALC and setting the attenuator to 20

dB.

Agilent Technologies 83640B — 08360125.pdf diagram (page 117)
Fig. 14 · 08360125.pdf

Figure 2-7. Maximum Lev eled P o w er T est Setup

A tten uator 20 dB HP/Agilent 8490D Option 020 (83640B/L and 83650B /L)

A tten uator 20 dB HP/Agilent 8493C Option 020 (83620B/22B/24B,

83623B/L, and 83630B /L)

2. T o ac hiev e p eak p ow er, initiate auto tracking on the syn thesizer as follo ws. T erminate

the RF OUTPUT with a go o d 50 ohm imp edance match suc h as a 20 dB attenuator or a

p ow er sensor (not necessary for syn thesizers with a step attenuator.) Press:

PO WER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MENU

W ait for the syn thesizer to complete auto tracking b efore con tinuing to the next step.

3. On the syn thesizer, press:

POWER LEVEL

Set the p ow er to 1 dB b elow the sp eci ed maximum lev eled p ow er for the

syn thesizer's full frequency range.

On (amb er light is on)

RF ON/OFF

CONT

SWEEP TIME

2-28 P erformance T ests

Maximum Lev eled P o w er

Show 13 more labels / short text on this page
Description and Procedure · Equipment · · Tracking Menu · Auto Track · · · · · · 0 · (minim um) · ms
Page 118 of the Agilent Technologies 83640B user manual — CONT
Page 118 · click to enlarge
Page 118
4. Increase the p ow er lev el un til the unlev eled status indicator comes on, then reduce the

p ow er lev el un til the indicator just go es o . (Pow er is lev eled).

5. Set the sweep to

and initiate sev eral sweeps. If necessary , reduce the p ow er lev el

SINGLE

un til the unlev eled status indicator do es not turn on while sweeping.

6. On the syn thesizer, press:

SWEEP TIME

0.5

Initiate sev eral sweeps. If necessary , reduce the p ow er lev el un til the unlev eled status

indicator do es not turn on while sweeping.

Record the p ow er lev el on the test record.

7. On the syn thesizer, press:

CONT

SWEEP TIME

8. F or the following syn thesizer mo dels, c hange the stop frequency to eac h of the frequencies

listed. Repeat steps 3 through 7 for eac h of these new frequency ranges.

83630B/L20.0 GHz
83640B/L26.5 GHz

20.0 GHz (Option 001 only)

83650B/L40.0 GHz

26.5 GHz

20.0 GHz (Option 001 only)

Ampli er/Multiplier Adjustmen ts

Ampli er/Filter Adjustmen ts

Mo dulator O set and Gain

In Case of Difficulty

Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

Agilent 8360 P erformance T ests

Maximum Lev eled P o w er

Show 11 more labels / short text on this page
· · sec · · · · · 0 · msec · (minim um) · Related Adjustments
Page 119 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 119 · click to enlarge
Page 119
7. External Lev eling

The syn thesizer is set up to externally lev el using a negative crystal detector. A DVM

measures the crystal detector DC output. A 1000 pf capacitor across the DVM input limits

the e ect of an y small signals generated b y the DVM. The syn thesizer p ow er lev el is set to the

minimum allo w able limit and then is increased un til the p ow er output is lev eled. The external

lev eling v oltage is then measured and compared to the minimum sp eci cation. Next, the

output p ow er is increased un til the maxim um lev eled p ow er is reac hed. The lev eling v oltage is

measured and compared to the sp eci cation.

The v oltage , not the p ow er lev el , is what is imp ortant in p erforming this test.

If y ou ha v e problems ac hieving the sp eci ed v oltages in this test, y ou may place a pad in the

RF path b efore the crystal detector or the coupling factors may b e c hanged to ac hiev e the

sp eci ed v oltages.

The syn thesizer m ust meet the p ow er atness sp eci cation b efore pro ceeding

with this test.

1. Preset all instrumen ts sho wn in Figure 2-8 and let them w arm up for at least one hour.
Agilent Technologies 83640B — 08360125.pdf diagram (page 119)
Fig. 15 · 08360125.pdf

Figure 2-8. External Lev eling T est Setup

Digital V oltmeter HP/Agilent 3456A

Crystal Detector HP/Agilent 33330E (83640B /L and 83650B/L)

Crystal Detector HP/Agilent 33330D (83620B/22B/24B,

Capacitor 1000 pf P/N 0160-4574

2. T o ac hiev e p eak p ow er, initiate auto tracking on the syn thesizer as follo ws: terminate the

RF output with a go o d 50 imp edance match suc h as a 10 dB attenuator or a p ow er

sensor (not necessary for syn thesizers with a step attenuator). Press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USER CAL

Auto Track (asterisk on)

W ait for the syn thesizer to complete auto tracking b efore con tinuing to the next step.

2-30 P erformance T ests

Show 7 more labels / short text on this page
Description and Procedure · Equipment · 83623B/L, and 83630B/L) · · · Tracking Menu · External Lev eling
Page 120 of the Agilent Technologies 83640B user manual — USER CAL
Page 120 · click to enlarge
Page 120
3. Connect the equipmen t as sho wn in Figure 2-8 .
4. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ALC

POWER LEVEL

5. Maximum accuracy for the HP/Agilent 3456A is obtained b y increasing the DVM display

resolution and the n um b er of in tegration cycles. On the HP/Agilent 3456A, set:

STORE

N CYC INT

STORE

N DIG DISP

Set the DVM to measure DC v olts.

6. Increase the syn thesizer p ow er lev el un til the LOW UNLVLED message turns o . (If it is

already out, leav e the p ow er lev el at 36 dBm.)

7. On the syn thesizer, press:

CW

Record the DVM reading on the test record and compare the reading to the sp eci cation.

If the UNLVLED message turns on b efore reac hing +4 dBm, set the stop

frequency to a low enough frequency where the sp eci ed output p ow er

is 4 dBm. If y ou don't get an unlev eled message, disconnect the BNC

tee temp orarily to make sure that the unlev eled message turns on. Then

reconnect the equipmen t and con tinue with the test.

8. Press

POWER LEVEL

. Increase the syn thesizer p ow er lev el un til the UNLVLED message turns

on. Decrease p ow er just un til the message go es out.

9. Record the DVM reading on the test record and compare the reading to the sp eci cation.

Ampli er/Multiplier Adjustmen ts

Ampli er/Filter Adjustmen ts

P ow er Flatness

In Case of Difficulty

1. Be sure y ou are externally lev eling with a negative crystal detector.
2. If the DVM readings are not within the sp eci cations, p erform an external detector

calibration (under the

USER CAL

k ey .)

3. Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

Agilent 8360 P erformance T ests

Show 28 more labels / short text on this page
· · Leveling Point ExtDet · Coupling Factor · 0 · dB · · · · 36 · · dBm · · · 100 · · · · 6 · · · · · 10 · GHz · · Related Adjustments · External Lev eling
Page 121 of the Agilent Technologies 83640B user manual — USER CAL
Page 121 · click to enlarge
Page 121
8. Spurious Signals (Harmonics & Subharmonics)

Use this pro cedure to measure the syn thesizer's harmonics and subharmonics ov er its en tire

frequency range. Harmonics are in teger m ultiples of the syn thesizer RF output frequency and

subharmonics are fractional m ultiples of the YO frequency . Subharmonics do not exist at all

frequencies. In low band, mixer spurs and fundamen tal feedthrough are the most signi cant

harmonically-related signals. The mixer spur/YO frequency relationship is as follows:

X (5.4 GHz xed oscillator) Y (YO frequency) = Mixer Spur F requency

where X and Y are in tegers.

Other harmonics are typically direct m ultiples of the YO frequency .

In this pro cedure, the syn thesizer is man ually swept ov er its frequency range while the

sp ectrum analyzer measures the harmonics and subharmonics in eac h frequency band. An y

harmonics or subharmonics that are within 5 dB of the sp eci cation are subsequen tly v eri ed

with a more accurate pro cedure.

1. T urn on the equipmen t sho wn in Figure 2-9 and Figure 2-10. Preset the instrumen ts and

let them w arm up for at least one hour.

On the HP/Agient 8566B, connect the CAL OUTPUT to the RF INPUT. Press

SHIFT

W

to calibrate the sp ectrum analyzer.

Agilent Technologies 83640B — SHIFT diagram (page 121)
Fig. 16

Figure 2-9. Spurious Signals (Harmonic) T est Setup < 20 GHz

Sp ectrum Analyzer HP/Agilent 8566B

A tten uator 10 dB HP/Agilent 8490D Option 010 (83640B/L

A tten uator 10 dB HP/Agilent 8493C Option 010 (83620B /22B/24B,

2. T o ac hiev e p eak p ow er, turn on RF p eaking. Set:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Peak RF Always (asterisk on).

USER CAL

2-32 P erformance T ests

Spurious Signals (Harmonics Subharmonics)

Show 9 more labels / short text on this page
Description and Procedure · · · Equipment · and 83650B /L) · 83623B/L, and 83630B/L) · · · Tracking Menu
Page 122 of the Agilent Technologies 83640B user manual — POWER LEVEL
Page 122 · click to enlarge
Page 122

Harmonic Measurement < 20 GHz

3. Connect the equipmen t as sho wn in Figure 2-9.
4. On the syn thesizer, press:

Set the maximum sp eci ed lev eled p ow er.

POWER LEVEL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SWEEP

MENU

5. On the sp ectrum analyzer, set the rst set of start and stop frequencies from T able 2-9.

Then set:

Reference Level 20 dBm
Scale Log5 dB/Division
Bandwidth Resolution3 MHz

Video Bandwidth 3 MHz

T able 2-9. Start and Stop Frequencies

2.0 7.0

7.0 13.5

13.5 20.0

1 Set to 2 GHz for 83622B/24B.

6. Manually sweep the syn thesizer across the frequency range while c hec king the sp ectrum

analyzer display for harmonics and subharmonics.

See T able 2-10 for the YO frequency ranges that corresp ond to the RF output frequencies.

T able 2-10.

Corresponding YO Frequency Ranges and RF Output Frequencies

5.41 to 7.4 N/A 0.01 to 2.0

2.0 to 7.0 1 2.0 to 7.0

3.5 to 6.75 2 7.0 to 13.5

4.5 to 6.67 3 13.5 to 20.0

7. Compare the amplitude of the harmonics to the sp eci cations listed in the test record. If

an y harmonic is within 5 dB of sp eci cation, make a more accurate measuremen t using

the \Harmonic/Subharmonic V eri cation Pro cedure < 20 GHz" that follo ws.

8. F or the 83620B , 83622B , and 83630B /L: On the test record, record the v alue of the w orst

case harmonic for RF output frequencies of 0.01 to 2.0 GHz.

Agilent 8360 P erformance T ests

Spurious Signals (Harmonics Subharmonics)

Show 12 more labels / short text on this page
· · Manual Sweep · Start F requency · Stop F requency · (GHz) · (GHz) · YO F requencies · Harmonic RF Output · (GHz) · F requencies · (GHz)
Page 123 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 123 · click to enlarge
Page 123

F or the 83624B , 83623B /L, 83640B /L, and 83650B /L and all for instrumen ts with Option

6. On the test record, record the v alue of the w orst case harmonic for RF output

frequencies of 10 to < 50 MHz and 50 MHz to 2.0 GHz.

9. Reset y our syn thesizer to the parameters given in step 5.

2-34 P erformance T ests

Spurious Signals (Harmonics Subharmonics)

Page 124 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 124 · click to enlarge
Page 124
10. F or eac h set of start and stop frequencies given in T able 2-9 man ually sweep the

syn thesizer across the frequency range and c hec k the sp ectrum analyzer display for

harmonics and subharmonics.

11. Compare the harmonics/subharmonics to the sp eci cations listed in the test record

and v erify an y that are within 5 dB of sp eci cation with the \Harmonic/Subharmonic

V eri cation Pro cedure < 20 GHz."

12. F or the 83620B /22B/24B and 83623B /L: Record, on the test record, the w orst case

harmonic for RF output frequencies of 2.0 and 20 GHz and the v alue of the w orst

case subharmonic for RF output frequencies of 7 to 20 GHz.

F or the 83630B /L, 83640B /L, and 83650B /L: Record, on the test record, the w orst case

subharmonic for RF output frequencies of 7 and 20 GHz. Note, for future use, the

w orst case harmonic for RF output frequencies of 2.0 to 20 GHz .

Harmonic Measurement 20 to 26.5 GHz

Disr e gar d if b eyond your synthesizer's c ap ability.

13. Connect the equipmen t as sho wn in Figure 2-10 using the HP/Agilent 11970K Mixer (20

to 26.5 GHz).

Agilent Technologies 83640B — 08360125.pdf diagram (page 124)
Fig. 17 · 08360125.pdf

Figure 2-10. Spurious Signals (Harmonics) T est Setup > 20 GHz

Sp ectrum Analyzer HP/Agilent 8566B

Micro w av e Ampli er HP/Agilent 11975A

Harmonic Mixer (20 to 26.5 GHz) HP/Agilent 11970K

Harmonic Mixer (26.5 to 40 GHz) HP/Agilent 11970A

Harmonic Mixer (40 to 50 GHz) HP/Agilent 11970Q

1 Firm w are revision 13.8.86 or later.

Agilent 8360 P erformance T ests

Spurious Signals (Harmonics Subharmonics)

Equipment

Page 125 of the Agilent Technologies 83640B user manual — 83650B/L
Page 125 · click to enlarge
Page 125
14. On the sp ectrum analyzer, select the external mixer for the 20 to 26.5 GHz frequency

range and set the reference lev el o set to comp ensate for the mixer's con v ersion loss at 23

GHz.

15. On the syn thesizer, press:

CW

16. Set the microw av e ampli er output p ow er lev el to +16 dBm.
17. On the sp ectrum analyzer, set:
Center Frequency23 GHz
Frequency Span1 MHz

Adjust the reference lev el to p osition the signal p eak at the top reference graticule. Use

this reference lev el for all harmonic measuremen ts.

18. On the sp ectrum analyzer, set:
Start Frequency20 GHz
Stop Frequency26.5 GHz
19. Manually sweep the syn thesizer across the frequency range while c hec king the sp ectrum

analyzer display for harmonics and subharmonics. See T able 2-11 for the YO frequency

ranges that corresp ond to the RF output frequencies.

Since an external mixer is used, use the sp ectrum analyzer signal iden tify feature to v erify

that susp ected signals are in the 20 to 26.5 GHz frequency range. The signals most likely

to app ear are the fth and sixth YO harmonics.

20. Compare the amplitude of the harmonics and subharmonics to the sp eci cations listed in

the test record. If an y harmonics or subharmonics are within 5 dB of the sp eci cation,

make a more accurate measuremen t using the \Harmonic/Subharmonic V eri cation

Pro cedure > 20 GHz" that follo ws. Note the w orst case harmonic and the

w orst case subharmonic from 20 to 26.5 GHz.

21. F or the 83630B /L only: Record, on the test record, the w orst case subharmonic from > 20

to 26.5 GHz.

F or the 83630B /L, 83640B /L, and 83650B /L only: F rom steps 12 and 20, determine the

w orst case harmonic from 2.0 to 26.5 GHz and record this v alue on the test record.

T able 2-11.

Corresponding YO Frequency Ranges and RF Output Frequencies

Mo del YO F requencies

5.0 to 6.625 4 4 20 to 26.5

83650B/L

83650B /L 4.41667 to 5.5667 3 6 26.5 to 33.4

83650B /L 5.5667 to 6.333 3 6 33.4 to 38

83650B /L 4.75 to 6.25 4 8 38 to 50

83640B /L 5.0 to 6.375 2 4 20 to 25.5

83640B /L 4.25 to 5.0 3 6 25.5 to 30

83640B /L 5.0 to 6.667 3 6 30 to 40

2-36 P erformance T ests

Spurious Signals (Harmonics Subharmonics)

Show 13 more labels / short text on this page
· · 23 · · GHz · YO Harmonic · YO Harmonic · RF Output · (GHz) · F rom Amp/Mult · at RF Output · F requencies (GHz) · 83630B /L and
Page 126 of the Agilent Technologies 83640B user manual — CW
Page 126 · click to enlarge
Page 126

Harmonic Measurement 26.5 to 40 GHz

Disr e gar d if b eyond your synthesizer's c ap ability.

22. Replace the HP/Agilent 11970K mixer with the HP/Agilent 11970A Mixer

(26.5 to 40 GHz).

23. On the sp ectrum analyzer, select the external mixer for the 26.5 to 40 GHz frequency

range and set the reference lev el o set to comp ensate for the mixer's con v ersion loss at 33

GHz.

24. On the syn thesizer, press:

CW

25. On the sp ectrum analyzer, set:
Center Frequency33 GHz
Frequency Span1 MHz

Adjust the reference lev el to p osition the signal p eak at the top reference graticule. Use

this reference lev el for all harmonic measuremen ts.

26. On the sp ectrum analyzer, set:
Start Frequency26.5 GHz
Stop Frequency40 GHz (83640B/L)
27. Manually sweep the syn thesizer across the frequency range while c hec king the sp ectrum

analyzer display for harmonics and subharmonics. See T able 2-11 for the YO frequency

ranges that corresp ond to the RF output frequencies.

Since an external mixer is used, use the sp ectrum analyzer signal iden tify feature to v erify

that susp ected signals are in the 26.5 to 40 GHz frequency range. The signals most likely

to app ear are the fth and sixth YO harmonics.

28. Compare the amplitude of the harmonics and subharmonics to the sp eci cations listed in

the test record. If an y harmonics or subharmonics are within 5 dB of the sp eci cation,

make a more accurate measuremen t using the \Harmonic/Subharmonic V eri cation

Pro cedure > 20 GHz" that follo ws. Note the w orst case harmonic and the

w orst case subharmonic from 26.5 to 40 GHz.

29. F or the 83640B /L only: F rom step 28, determine the w orst case harmonic from 26.5 to

40 GHz and record this v alue on the test record.

F or the 83640B /L/ and 83650B /L only: F rom steps 20 and 28, determine the w orst case

subharmonic from > 20 and 40 GHz and record this v alue on the test record.

Harmonic Measurement 38 to 50 GHz

Disr e gar d if b eyond your synthesizer's c ap ability.

30. Replace the HP/Agilent 11970A mixer with the HP/Agilent 11970Q Mixer (33 to 50

GHz).

31. On the sp ectrum analyzer, select external mixer for the 33 to 50 GHz frequency range and

set the reference lev el o set to comp ensate for the con v ersion loss at 40 GHz.

32. On the syn thesizer, press:

CW

Agilent 8360 P erformance T ests

Spurious Signals (Harmonics Subharmonics)

Show 11 more labels / short text on this page
· · 33 · · GHz · 38 GHz (83650B/L) · · · 40 · GHz ·
Page 127 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 127 · click to enlarge
Page 127
33. On the sp ectrum analyzer, set:
Center Frequency40 GHz
Frequency Span1 MHz

Adjust the reference lev el to p osition the signal p eak at the top reference graticule. Use

this reference lev el for all harmonic measuremen ts.

34. On the sp ectrum analyzer, set:
Start Frequency38 GHz
Stop Frequency50 GHz
35. Manually sweep the syn thesizer across the frequency range while c hec king the sp ectrum

analyzer display for harmonics and subharmonics. See T able 2-11 for the YO frequency

ranges that corresp ond to the RF output frequencies.

Since an external mixer is used, use the sp ectrum analyzer signal iden tify feature to v erify

that susp ected signals are in the 38 to 50 GHz frequency range. The signals most likely to

app ear are the fth and sixth YO harmonics.

36. Compare the amplitude of the harmonics and subharmonics to the sp eci cations listed in

the test record. If an y harmonics or subharmonics are within 5 dB of the sp eci cation,

make a more accurate measuremen t using the \Harmonic/Subharmonic V eri cation

Pro cedure > 20 GHz" that follo ws. Note the w orst case harmonic and the

w orst case subharmonic from 38 to 50 GHz.

37. F or the 83650B /L only: F rom steps 28 and 36, determine the w orst case harmonic from

26.5 to 50 GHz and record this v alue on the test record. F rom step 36, determine the

w orst case subharmonic from > 40 and 50 GHz and record this v alue on the test record.

Harmonic/Subharmonic V erification Procedure < 20 GHz

1. Set the syn thesizer to the RF output frequency that corresp onds to the harmonic or

subharmonic to b e measured.

2-38 P erformance T ests

Spurious Signals (Harmonics Subharmonics)

Page 128 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 128 · click to enlarge
Page 128
2. On the sp ectrum analyzer, set:
FrequencySame frequency as the synthesizer
Span1 MHz
Reference Level15 dBm
Scale Log5 dB/Division
Bandwidth Resolution10 kHz
Video Bandwidth30 kHz
3. Measure the syn thesizer RF output amplitude with the sp ectrum analyzer marker.
4. Set the sp ectrum analyzer to the harmonic or subharmonic frequency to b e measured.

Measure the signal lev el with the sp ectrum analyzer marker. Change the reference lev el as

necessary .

5. Calculate the harmonic/subharmonic lev el, where the harmonic/subharmonic is less than

the carrier, as follo ws:

Harmonic/Subharmonic amplitude (dBc) = [Syn thesizer RF output amplitude (dBm)

F or example:

RF Output = +10 dBm

Harmonic = 60 dBm

Harmonic (dBc) = [RF Out (Harmonic)]

= [10 (60 dBm)]

= [10 + 60 dBm]

= 70 dBc

Harmonic/Subharmonic V erification Procedure > 20 GHz

F or accurate measuremen ts, the syn thesizer m ust meet its p ow er atness

sp eci cation.

1. Note the syn thesizer CW frequency that pro duces the susp ect harmonic/subharmonic.
2. Set the syn thesizer to the same RF output frequency as the harmonic/subharmonic to b e

measured.

Agilent 8360 P erformance T ests

Spurious Signals (Harmonics Subharmonics)

(Harmonic/Subharmonic amplitude (dBm))]

Page 129 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 129 · click to enlarge
Page 129
3. On the sp ectrum analyzer, set:
FrequencySame frequency as the synthesizer
Span1 MHz
Reference Level15 dBm
Scale Log5 dB/Division
Bandwidth Resolution10 kHz
Video Bandwidth30 kHz
4. Measure the syn thesizer RF output amplitude with the sp ectrum analyzer marker.
5. Set the syn thesizer to the CW frequency noted in step 1.
6. Measure the signal lev el with the sp ectrum analyzer marker. Change the reference lev el as

necessary .

7. Calculate the harmonic/subharmonic lev el where the harmonic/subharmonic is less than

the carrier, as follo ws:

Harmonic/Subharmonic Amplitude (dBc) = [Syn thesizer RF output amplitude (dBm)

F or example:

RF Output = +10 dBm

Harmonic = 60 dBm

Harmonic (dBc) = [RF Out (Harmonic)]

= [10 (60 dBm)]

= [10 + 60 dBm]

= 70 dBc

In Case of Difficulty

Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

2-40 P erformance T ests

Spurious Signals (Harmonics Subharmonics)

Harmonic/Subharmonic amplitude (dBm)] · Related Adjustments · None

Page 130 of the Agilent Technologies 83640B user manual — POWER LEVEL
Page 130 · click to enlarge
Page 130
9. Spurious Signals (Non-Harmonic)

Use this pro cedure to measure kno wn, xed, o set spurs that are generated in the frequency

syn thesis section of the syn thesizer. The syn thesizer is set to v arious CW frequencies where

these spurious signals will most likely o ccur. Then the sp ectrum analyzer is tuned to the spur

frequencies to measure their lev els.

1. Preset the instrumen ts sho wn in Figure 2-11 and let them w arm up for at least one hour.

On the HP/Agilent 8566B, connect the CAL OUTPUT to the RF INPUT. Press

SHIFT

W

to calibrate the sp ectrum analyzer.

Agilent Technologies 83640B — SHIFT diagram (page 130)
Fig. 18

Figure 2-11. Spurious Signals (Non-Harmonic) T est Setup

Sp ectrum Analyzer HP/Agilent 8566B

2. T o ac hiev e p eak p ow er, turn on RF p eaking. Set:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USER CAL

Peak RF Always (asterisk on)

3. Connect the equipmen t as sho wn in Figure 2-11.
4. On the syn thesizer, press:

CW

19.765

POWER LEVEL

Set the maximum sp eci ed lev eled p ow er.

5. On the sp ectrum analyzer, set:
Center Frequency19.765 GHz
Frequency Span500 kHz
Reference Level10 dBm
Scale Log5 dB/Division
Resolution BandwidthAuto
Sweep TimeAuto
MarkerSet to 19.765 GHz

Agilent 8360 P erformance T ests

Show 14 more labels / short text on this page
Description and Procedure · · · Equipment · · · Tracking Menu · Fixed Spurious Signals · · · GHz · · · Spurious Signals (Non-Harmonic)
Page 131 of the Agilent Technologies 83640B user manual — N/A 125 19,765,125,000
Page 131 · click to enlarge
Page 131
6. On the sp ectrum analyzer, set the marker to the highest p eak and then set the marker

to cen ter frequency . Decrease the frequency span to 100 Hz k eeping the signal cen tered

on the display and then rep eat the marker p eak searc h and set the marker to cen ter

frequency .

7. Note the signal amplitude on the sp ectrum analyzer. This is the carrier amplitude to

whic h the spurs are referenced.

8. Use the following pro cedure to determine the actual frequency of the spurs in T able 2-12 :

a.

On the syn thesizer, press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Select Cal .

SERVICE

b. Use the RPG knob to select the rst calibration constant in T able 2-12 (#18).

c. T o the v alue of the sp ectrum analyzer cen ter frequency (19.765 GHz), add the v alue of

the calibration constant (or twice the v alue, as indicated in T able 2-12) and the v alue

from the second column of T able 2-12 (20 kHz). Record this v alue (in Hz) in the last

column of T able 2-12 .

For example:

19.765 GHz (Spectrum analyzer center frequency)

+ 338 kHz (Possible value of calibration constant #18

+ 20 kHz (Value from second column of the table)

= 19,765,358,000 Hz (Record this result in the last column

of the table.)

Repeat this pro cedure for the remaining calibration constants in T able 2-12. The

last tw o actual spur frequencies ha v e already b een calculated and are not related to

calibration constants.

T able 2-12. Determining the Actual Spur Frequencies

Calibration Constan t #18 20

2(Calibration Constan t #18) 40

Calibration Constan t #19 60

2(Calibration Constan t #19) 120

N/A 125 19,765,125,000

N/A 500 19,765,500,000

9. On the sp ectrum analyzer, set the cen ter frequency to the rst \Actual Spur F requency"

from T able 2-12 . Then set:

Reference Level 50 dBm
Video AveragingOn 100 samples
MarkerSame as spectrum analyzer frequency

2-42 P erformance T ests

Show 12 more labels / short text on this page
dBm Carrier Amplitude · · Adjust Menu · Calib Menu · --------------------------------------------------------- ------------- · Calibration Constan t · Additional V alue · Actual Spur · (Hz) · (kHz) · F requency (Hz) · Spurious Signals (Non-Harmonic)
Page 132 of the Agilent Technologies 83640B user manual — CW
Page 132 · click to enlarge
Page 132
10. Lo cate the spur corresp onding to the sp ectrum analyzer frequency (see T able 2-12) and

use the marker to measure its amplitude. If the spur is in the noise lev el, use the noise

lev el amplitude (this gives a w orst case v alue).

dBm Spur Absolute Amplitude

11. Calculate the spur lev el in dBc as follows:

Carrier Amplitude (dBm) Spur Absolute Amplitude (dBm) = Spur Lev el (dBc)

Record the result on the test record. Compare the result to the sp eci cation.

12. Repeat steps 9 through 11 for eac h of the spurs and sp ectrum analyzer frequencies in

T able 2-12.

Lo w Band Offset Spurious Signals

Disr e gar d if b eyond your synthesizer's c ap ability.

13. On the syn thesizer, press:

CW

POWER LEVEL

14. On the sp ectrum analyzer, set:
Center Frequency1 GHz
Frequency Span500 Hz
Reference Level0 dBm
Resolution BandwidthAuto
Video BandwidthAuto
Marker1 GHz
Video AveragingOff
15. Use the marker to measure the syn thesizer's RF output amplitude at 1 GHz. This is the

carrier amplitude to whic h the spurs are referenced.

dBm RF output at 1 GHz

16. Set the sp ectrum analyzer cen ter frequency to eac h of the frequencies in T able 2-13 . Use

the marker to measure the spurs (c hange the reference lev el as necessary).

Calculate the spur lev el in dBc for eac h of the spurs as follows:

Spur Lev el (dBc) = [RF Output at 1 GHz (dBm) (Spur Lev el)]

Record the spur lev el in dBc on the test record. Compare the spur lev el to the

sp eci cation.

T able 2-13. Lo w Band Spurious Signals

Sp ectrum Analyzer F requency Spur

100 MHz 100 MHz Fixed Spur

900 MHz 100 MHz O set Spur

6.4 GHz LO F eedthrough Spur

Agilent 8360 P erformance T ests

Show 11 more labels / short text on this page
· · 1 · · GHz · · · 0 · · dBm · Spurious Signals (Non-Harmonic)
Page 133 of the Agilent Technologies 83640B user manual — CW
Page 133 · click to enlarge
Page 133

Lo w Band Mixer Spurs

17. On the syn thesizer, press:

1.6

CW

18. On the sp ectrum analyzer, set:
Start Frequency1.4 GHz
Stop Frequency1.8 GHz
Reference Level0 dBm
Resolution Bandwidth300 kHz
Video Bandwidth100 kHz
Sweep TimeAuto
Scale Log10 dB/Division
MarkerNormal Peak Search MKR-->REF LVL
19. Set the syn thesizer to CW 1.9999 GHz and tune from 1.9999 GHz to 1.8 GHz. This

results in the most dominant mixing spur moving from 1.4 to 1.8 GHz. The spur should

b e at the left-hand edge of the display (1.4 GHz). It should then trav el to the right as

y ou tune do wn. If y ou can't see it, increase p ow er un til y ou can iden tify the spur. Then

reset the p ow er lev el to 0 dBm (the p erformance sp eci cation is v alid for 0 dBm). Y ou are

nished tuning when the spur merges with the 2 GHz carrier.

20. Iden tify the w orst p oint (where the spur has the greatest p ow er lev el), or if the spur is lost

in the noise o or use that v alue, and record the di erence b et w een the reference lev el and

spurious signal amplitude on the test record.

F ractional-N Reference and API Spurs

In Case of Difficulty

Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

2-44 P erformance T ests

Show 6 more labels / short text on this page
· · · GHz · Related Adjustments · Spurious Signals (Non-Harmonic)
Page 134 of the Agilent Technologies 83640B user manual — POWER LEVEL
Page 134 · click to enlarge
Page 134
10. Spurious Signals (Line-Related)

Use this pro cedure to measure line-related spurs. The syn thesizer is placed at a CW frequency

where the syn thesizer is most sensitive to line-related spurs. The sp ectrum analyzer measures

the RF output o set b y the harmonics of the line p ow er frequency (60 Hz or 50 Hz). T o

eliminate measuring the sp ectrum analyzer line-related spurs, an In v ertron is used to op erate

the sp ectrum analyzer at a di eren t line frequency so that its own spurs will not a ect the

measuremen t.

The sp ectrum analyzer m ust ha v e the capabilit y of a 10 Hz resolution bandwidth. A wider

resolution bandwidth results in the syn thesizer's phase noise masking the spur.

1. Set the In v ertron for a line frequency of 55 Hz and for normal op erating v oltage.
2. Connect the equipmen t as sho wn in Figure 2-12. Preset the instrumen ts and let them

w arm up for at least one hour.

Agilent Technologies 83640B — 08360125.pdf diagram (page 134)
Fig. 19 · 08360125.pdf

Figure 2-12. Spurious signals (Line-Related) T est Setup

Sp ectrum Analyzer HP/Agilent 8566B

In v ertron California Instrumen ts 501TC

3. T o ac hiev e p eak p ow er, turn on RF p eaking. Set:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Peak RF Always (asterisk on)

USER CAL

4. On the syn thesizer, press:

CW

6.99

POWER LEVEL

5. On the sp ectrum analyzer, set:
Center Frequency6.99 GHz
Frequency Span2 kHz
Reference Level10 dBm
Scale Log5 dB/Division
Resolution Bandwidth300 Hz
Sweep TimeAuto

Agilent 8360 P erformance T ests

Show 15 more labels / short text on this page
Description and Procedure · Equipment · · · Tracking Menu · · · GHz · · · · 0 · · dBm · Spurious Signals (Line-Related)
Page 135 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 135 · click to enlarge
Page 135
MarkerSet to 6.99 GHz
6. On the sp ectrum analyzer, set the marker to the highest p eak and note the frequency

di erence of the marker from the cen ter frequency . Calibrate the sp ectrum analyzer

frequency o set so that the signal p eak is in the cen ter of the display .

7. Note the signal amplitude on the sp ectrum analyzer. This is the carrier amplitude to

whic h the spurs are referenced.

This pro cedure is written for a line frequency of 60 Hz. F or a 50 Hz line

frequency , lo ok for spurs at frequencies of 50 Hz times the harmonics in

T able 2-14, and set the sp ectrum analyzer frequencies accordingly .

8. On the sp ectrum analyzer, set the cen ter frequency to the rst spur frequency from

T able 2-14. Then set:

Reference Level 50 dBm
Frequency Span100 Hz
Resolution Bandwidth10 Hz
MarkerSame as spectrum analyzer frequency
Video AveragingOn 100 samples
9. Lo cate the spur corresp onding to the sp ectrum analyzer frequency (see T able 2-14 ) and

use the marker to measure its amplitude. If the spur is in the noise lev el, use the noise

lev el amplitude (this gives a w orst case v alue).

dBm Spur Absolute Amplitude

10. Calculate the spur lev el in dBc as follows:

Spur Lev el (dBc) = [Carrier Amplitude (dBm) (Spur Lev el)]

Record the result on the test record. Compare the result to the sp eci cation.

The o dd harmonics will normally b e higher than the ev en harmonics.

11. Repeat steps 8 through 10 for eac h of the spurs and sp ectrum analyzer frequencies in

T able 2-14. Change the sp ectrum analyzer reference lev el as indicated in the table.

T able 2-14.

Spectrum Analyzer Frequencies and Line Spur Frequencies and Harmonics

2 120 6,990,000,120 50

3 180 6,990,000,180 50

4 240 6,990,000,240 50

12. F or syn thesizers capable of frequencies do wn to 10 MHz, set the syn thesizer and sp ectrum

analyzer to 1.9 GHz and rep eat steps 6 through 11 for this frequency .

2-46 P erformance T ests

Show 10 more labels / short text on this page
dBm Carrier Amplitude · Line Spur · Line Spur · Sp ectrum Analyzer · Reference Lev el · Harmonic · F requency (Hz) · F requency (Hz) · (dBm) · Spurious Signals (Line-Related)
Page 136 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 136 · click to enlarge
Page 136

In Case of Difficulty

Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

Agilent 8360 P erformance T ests

Related Adjustments · None · Spurious Signals (Line-Related)

Page 137 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 137 · click to enlarge
Page 137
11. Single Sideband Phase Noise

This pro cedure uses a phase noise measuremen t system. It is the fastest and most accurate

metho d for measuring phase noise.

1. Connect the equipmen t as sho wn in Figure 2-13 . Preset the equipmen t and let them w arm

up for at least one hour.

2-48 P erformance T ests

Single Sideband Phase Noise

Description and Procedure

Page 138 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 138 · click to enlarge
Page 138
Agilent Technologies 83640B — 08360125.pdf diagram (page 138)
Fig. 20 · 08360125.pdf

Figure 2-13. Single Sideband Phase Noise T est Setup

Phase Noise Measuremen t System HP/Agilent 3048A

Agilent 8360 P erformance T ests

Single Sideband Phase Noise

Equipment

Page 139 of the Agilent Technologies 83640B user manual — POWER LEVEL
Page 139 · click to enlarge
Page 139
2. Load the measuremen t softw are for the phase noise measuremen t system.
3. On the syn thesizer, press:

CW

6.99

POWER LEVEL

4. On the phase noise measuremen t system, set:
Measurement Typephase locked
Start Offset Freq100 Hz
Stop Freq2 MHz
Minimum Averages40
Carrier Frequency6.99 GHz
Det Input Freq50 MHz
DUTuser's srce, man
Ref Source8662A, SYS, VCO, EFC
Ext Timebasenot in use
Down Converter1179C, SYS
HP 11848A LNAout

T o av oid en tering these parameters eac h time y ou p erform this test, create a

le on the phase noise measuremen t system con taining these parameters.

With these parameters en tered, the softw are automatically sets the remaining parameters.

T able 2-15 sho ws the complete set of parameters.

T able 2-15. Phase Noise Measurement System P arameters

Measurement T yp ephase lo cked K V CO Metho d: measured
Start O set F req100 Hz V CO T une Constan t: 82.33 Hz/V
Stop F req2 MHz
Minim um Av erages40
Lo op SuppressionV eri ed
Closed PLL BW150.5 Hz
Carrier F requency6.99 GHz P eak T uning Range: 725.1 Hz
Det Input F req50 MHz Assumed P ole: 37.9 kHz
Entered K V CO28 Hz/V DUT: user's srce, man
Center V oltage0 V Ref Source: 8662A, SYS, V CO, EF C
T une-v oltage Range 10 V Ext Timebase: not in use
Phase Detector5 to 1600 MHz Down Converter: 11792C, SYS
K Detector Metho dmeasured HP 11848A LNA: Out
Detector Constan t458.8 mV/Rad
5. F ollo w the instructions on the con troller to make the phase noise measuremen t.
6. The dynamic signal analyzer has tw o traces. The sine w av e sho ws the b eatnote frequency

in time domain (like an oscilloscop e). The other trace is in the frequency domain (like a

sp ectrum analyzer).

2-50 P erformance T ests

Single Sideband Phase Noise

Show 9 more labels / short text on this page
· · GHz · · · · 0 · dBm ·
Page 140 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 140 · click to enlarge
Page 140

A t the message VERIFY BEATNOTE

w av e frequency with the frequency domain signal near the left edge (0 Hz).

7. When the measuremen t is complete record the results on the test record and compare them

to the sp eci cation.

8. Repeat steps 3 through 6 for the frequencies in T able 2-16 .

T able 2-16. Frequency Setting for Phase Noise Measurements

Phase Noise Measurement System

1.97 310

18.0 18.0 720

1 Disregard if b eyond your synthesizer's capabilit y .

In Case of Difficulty

1. Line spurs may b e presen t in the trace whic h exceed the phase noise sp eci cation. They

should b e ignored.

2. Be sure that the signal generator frequency (detector input) is tuned close enough that a

zero b eat can b e found.

3. Note where the problem o ccurs (carrier frequency and o set frequency from the carrier)

and refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

Agilent 8360 P erformance T ests

Single Sideband Phase Noise

Show 9 more labels / short text on this page
Synthesizer · F requency · (GHz) · Carrier F requency · Detector Input · (GHz) · (MHz) · Related Adjustments · Sampler Assem bly
Page 141 of the Agilent Technologies 83640B user manual — W
Page 141 · click to enlarge
Page 141
12. Pulse Modulation On/Off Ratio

(F or B-Series Only)

Using a sp ectrum analyzer, the syn thesizer's CW RF output p ow er is measured b oth with

pulse on and with pulse o . The di erence in p ow er is the pulse on/o ratio.

1. Connect the equipmen t as sho wn in Figure 2-14. Preset all instrumen ts and let them

w arm up for at least one hour.

On the HP/Agilent 8556B, connect the CAL OUTPUT to the RF INPUT. Press

SHIFT

to calibrate the sp ectrum analyzer.

W

Agilent Technologies 83640B — W diagram (page 141)
Fig. 21

Figure 2-14. Pulse Modulation On/Off Ratio T est Setup

Sp ectrum Analyzer HP/Agilent 8566B

2-52 P erformance T ests

Pulse Modulation On/Off Ratio

Show 4 more labels / short text on this page
Description and Procedure · · · Equipment
Page 142 of the Agilent Technologies 83640B user manual — MOD
Page 142 · click to enlarge
Page 142
2. On the syn thesizer, press:

CW

The rst cen ter frequency in T able 2-17.

F or the 83623B/24B onlySet the p ow er lev el to 0 dBm b y pressing:

POWER LEVEL

F or all other syn thesizers: Set the p ow er lev el to 9.9 dBm b y pressing:

9.9

POWER LEVEL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Then, for all syn thesizersPress: FREQUENCY

MENU

Up/Dn Size CW

3. On the sp ectrum analyzer, set:
Center FrequencyThe same frequency set in step 2
Frequency Span100 Hz
Reference Level0 dBm
MarkerOn
Center Frequency Step Size50 MHz

T able 2-17. Pulse On/Off Center Frequencies

Cen ter F requency (GHz) Pulse On/O Ratio

1 Disregard if b ey ond your synthesizer's capability .

4. On the sp ectrum analyzer:

a. T une cen ter frequency to cen ter the signal on the display .

b. Set the marker to highest p eak.

c. Note the marker amplitude.

5. On the syn thesizer, turn on the external pulse. Press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Pulse On/Off Extrnl (asterisk on)

MOD

or, for syn thesizers with Option 002, press:

MOD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Pulse On/Off Ext (asterisk on)

6. On the sp ectrum analyzer set the reference lev el to 70 dBm.
7. Set the marker to the highest p eak and note the maxim um marker amplitude.
8. The di erence b et w een the marker v alue in step 4 and the marker v alue in step 7 is the

pulse on/o ratio. Record this v alue in T able 2-17.

9. On the sp ectrum analyzer, set the reference lev el to 0 dBm.
10. On the syn thesizer, turn o external pulse:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MOD

Pulse On/Off Extrnl (asterisk o )

or, for syn thesizers with Option 002, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Pulse On/Off Ext (asterisk o )

MOD

11. Repeat steps 2 through 10 for the remaining syn thesizer and sp ectrum analyzer

frequencies in T able 2-17 with the following c hanges to step 7 when testing 6.0 GHz:

Agilent 8360 P erformance T ests

Pulse Modulation On/Off Ratio

Show 25 more labels / short text on this page
· · · 0 · dBm · · · · · · dBm · · 50 · MHz · · · · · · Pulse Menu · · · · · Pulse Menu
Page 143 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 143 · click to enlarge
Page 143

F or a CW F requency of 6.0 GHz:

a. Step the syn thesizer and sp ectrum analyzer in 50 MHz steps from 6.0 to 7.0 GHz.

b. Note the frequency of the highest amplitude signal. Set the syn thesizer and sp ectrum

analyzer to this frequency .

c. Set the marker to the highest p eak and note the maximum marker amplitude.

12. Record the w orst case v alue from T able 2-17 on the test record.

In Case of Difficulty

1. A failure can b e caused b y a spur. Mov e the cen ter frequency of b oth the sp ectrum

analyzer and the syn thesizer b y 1 kHz, and retest at that p oint. If the on/o ratio is bad,

it will b e bad ov er greater than a 100 kHz region.

2. Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

2-54 P erformance T ests

Pulse Modulation On/Off Ratio

Related Adjustments · None

Page 144 of the Agilent Technologies 83640B user manual — CW
Page 144 · click to enlarge
Page 144
13. Pulse P erformance

(F or B-Series Only)

In this pro cedure, the syn thesizer's RF output is display ed directly on a high-bandwidth

digitizing oscilloscop e where the rise and fall times are measured and a comparison of pulsed

and CW p ow er lev el accuracy is made. P ow er lev el accuracy is c hec k ed at a 10 Hz pulse

rep etition rate to v erify its op eration at the slow est sp eci ed rate.

Rise and F all Times

1. Connect the equipmen t as sho wn in Figure 2-15. Preset the instrumen ts and let them

w arm up for at least one hour.

Agilent Technologies 83640B — 08360125.pdf diagram (page 144)
Fig. 22 · 08360125.pdf

Figure 2-15. Pulse P erformance T est Setup

Digitizing Oscilloscop e HP/Agilent 54124T

Pulse Generator HP/Agilent 8112A

A tten uator 10 dB HP/Agilent 8490D Option 010 (83640B/50B)

A tten uator 10 dB HP/Agilent 8493C Option 010 (83620B/22B/23B/24B/30B)

A tten uator 20 dB HP/Agilent 8490D Option 020 (83640B/50B)

A tten uator 20 dB HP/Agilent 8493C Option 020 (83620B/22B/23B/24B/30B)

2. On the syn thesizer, press:

First syn thesizer frequency in T able 2-18.

CW

POWER LEVEL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ALC

Leveling Point Internal (asterisk on)

Agilent 8360 P erformance T ests

Show 11 more labels / short text on this page
Description and Procedure · Equipment · · · · 0 · · dBm · · · Pulse P erformance
Page 145 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 145 · click to enlarge
Page 145

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Leveling Mode Search (asterisk on)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USER CAL

Peak RF Always (asterisk on)

T able 2-18. Synthesizer Frequencies

Risetime F alltime Lev el

1 Disregard if b eyond your synthesizer's capability .

The CW frequencies c hec k ed provide a minimum v eri cation of pulse

p erformance. If pulse p erformance at a di eren t frequency is imp ortant,

rep eat the test at that CW frequency .

3. On the pulse generator, set:
Pulse Width50 ns
Pulse Period10 s (100 kHz)
Pulse Level5 V
DisableLED off (enables pulse generator)
4. On the oscilloscop e, set:

Channel 3:

Volts/Division 15 mV (Readjust as required)

Timebase:

Delay Reference At center

Trigger:

Trigger Level 1.2 V

Display:

2-56 P erformance T ests

Show 14 more labels / short text on this page
· · Tracking Menu · Synthesizer · F requencies (GHz) · Accuracy · Ext Probe 10 · Display On · Offset 0 V · Time/Division 5 ns · Delay 100 ns · Trigger Slope Positive · Probe Atten 20 · Pulse P erformance
Page 146 of the Agilent Technologies 83640B user manual — MOD
Page 146 · click to enlarge
Page 146

Display Time 1 sec

On the oscilloscop e, turn the w av eform math function on. Then use the

w av eform math maxim um function to determine the pulse en v elop e when

making the follo wing measuremen ts.

5. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MOD

Pulse On/Off Extrnl (asterisk on)

or, for syn thesizers with Option 002, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Pulse On/Off Ext (asterisk on)

MOD

6. On the oscilloscop e:

a. Adjust the timebase dela y to p osition the rising edge of the pulsed RF near the cen ter

of the display .

b. Adjust the c hannel 3 v olts/division and o set to obtain a 5 division signal lev el

b et w een the RF p ow er o and the RF p ow er on. (See Figure 2-16 .)

c. Adjust the c hannel 3 o set to mov e the RF p ow er o line one-half division b elow the

nearest horizontal graticule (making the horizontal graticule cross the w av eform at the

10% p oint.)

d. Adjust the oscilloscop e timebase to widen the pulse en v elop e so that the measuremen t

can b e made with greater accuracy . (See Figure 2-17 .)

e. Use the oscilloscop e's delta T function to measure the time di erence b et w een the 10%

and 90% risetime p oints on the en v elop e. (The 10% p oint of the risetime is where the

pulse en v elop e crosses the graticule. The 90% p oint is 4 divisions up.) Record this

v alue in T able 2-18.

Agilent Technologies 83640B — MOD diagram (page 146)
Fig. 23

Figure 2-16. Pulse Env elope Display ed Ov er 5 Divisions

Agilent 8360 P erformance T ests

Show 8 more labels / short text on this page
Display Mode Persistence · Screen Single · · · · · Pulse Menu · Pulse P erformance
Page 147 of the Agilent Technologies 83640B user manual — MOD
Page 147 · click to enlarge
Page 147
7. Adjust the timebase dela y to p osition the falling edge of the pulse near the cen ter of the

oscilloscop e display .

8. Measure the falltime the same w ay as y ou measured the risetime (see step 6). Record this

v alue in T able 2-18 .

9. Repeat steps 5 through 8 at eac h syn thesizer frequency in T able 2-18 .
10. Record eac h w orst case risetime and falltime v alue from T able 2-18 on the test record.
Agilent Technologies 83640B — 08360125.pdf diagram (page 147)
Fig. 24 · 08360125.pdf

Figure 2-17. Widened Pulse Env elope

Pulse Lev eling Accuracy

11. On the pulse generator, set:
Pulse Width1 s
Pulse Period1 ms
Pulse Level5 V
12. On the syn thesizer, press:

First syn thesizer frequency in T able 2-18

CW

POWER LEVEL

ALC

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MOD

Pulse On/Off Extrnl (asterisk on)

or, for syn thesizers with Option 002, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Pulse On/Off Ext (asterisk on)

MOD

2-58 P erformance T ests

Show 15 more labels / short text on this page
· · · 0 · dBm · · · · Leveling Mode Normal · · · · · Pulse Menu · Pulse P erformance
Page 148 of the Agilent Technologies 83640B user manual — MOD
Page 148 · click to enlarge
Page 148
13. On the oscilloscop e, set:

Channel 3:

Timebase:

Delay Reference At left

Trigger:

Trigger Level 50 mV

Increase the oscilloscop e sensitivit y to 5 mV/division. Adjust the dela y and timebase so

that the pulse w av eform takes up the en tire display . Adjust the oscilloscop e c hannel 3

o set to p osition the top of the pulse en v elop e near the cen ter graticule.

14. On the syn thesizer, turn o external pulse and set the p ow er lev el to 0.3 dBm. Press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Pulse On/Off Ext (asterisk o )

POWER LEVEL

.3

15. Use the oscilloscop e's w av eform math maximum function to determine the p eak p ow er

lev el (if the w av eform math function w as previously on, clear the display b efore executing

this step.) Use the oscilloscop e's delta V feature and p osition marker 1 at the cen ter of

the p eak p ow er as display ed b y the math w av eform (the width of the w av eform is caused

b y noise in the system.) Lea v e the marker at that p osition.

16. Set the syn thesizer's p ow er lev el to +0.3 dBm. Press:

POWER LEVEL

17. Clear the oscilloscop e display and use the oscilloscop e's w av eform math maximum feature

to determine the p eak p ow er lev el. Use the oscilloscop e's delta V feature and p osition

marker 2 at the cen ter of the p eak p ow er as display ed b y the math w av eform.

If the di erence b et w een marker 1 and marker 2 is less than 1 division, rep eat steps 13

through 17 with c hannel 3 v olts/division set to a more sensitive scale.

18. On the syn thesizer, press:

POWER LEVEL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Pulse On/Off Extrnl (asterisk on)

MOD

19. Observ e the p eak pulse p ow er. It should stay within the tw o delta V markers set in steps

16 and 18. Record the \Pass" or \F ail" in the Lev el Accuracy column in T able 2-18.

20. Repeat steps 12 through 19 at eac h syn thesizer frequency in T able 2-18.
21. If all results in T able 2-18 are Pass , record \Pass" on the test record. If an y results are

F ail , record \F ail" on the test record.

In Case of Difficulty

Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

Agilent 8360 P erformance T ests

Show 22 more labels / short text on this page
Volts/Division 30 mV · Time/Division 100 ns · · · · dBm · · · · · dBm · · · · 0 · dBm · · · · Related Adjustments · None · Pulse P erformance
Page 149 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 149 · click to enlarge
Page 149
14. Pulse P erformance (Alternate Procedure)

(F or B-Series Only)

In this pro cedure, the syn thesizer's RF output is do wncon v erted in frequency so that an

oscilloscop e can measure the rise and fall times and make a comparison of pulsed and CW

p ow er lev el accuracy . F or measuremen ts ab ov e 20 GHz RF output frequency , a sp ectrum

analyzer with external mixer is used to do wncon v ert to a 321.4 MHz RF. P ow er lev el accuracy

is c hec k ed at a 10 Hz pulse rep etition rate to v erify its op eration at the slow est sp eci ed

rate. The p ow er ampli er ensures a sucien t signal lev el in to the oscilloscop e to make the

measuremen t.

1. T urn on the equipmen t sho wn in Figure 2-18 and Figure 2-20. Preset the instrumen ts and

let them w arm up for at least one hour.

Rise and F all Times < 20 GHz

This pro cedure is not v alid for fast pulse option (Option 006).

2. Connect the equipmen t as sho wn in Figure 2-18.

2-60 P erformance T ests

Pulse P erformance (Alternate)

Description and Procedure

Page 150 of the Agilent Technologies 83640B user manual — (83620B/22B/23B/24B/30B)
Page 150 · click to enlarge
Page 150
Agilent Technologies 83640B — 08360125.pdf diagram (page 150)
Fig. 25 · 08360125.pdf

Figure 2-18. Pulse P erformance T est Setup < 20 GHz

(Syn thesized Sw eeper) 83620A/B

Digitizing Oscilloscop e HP/Agilent 54111D

Preampli er HP/Agilent 8447F

A tten uator 6 dB HP/Agilent 8493C Option 006

A tten uator 10 dB HP/Agilent 8490D Option 010

(83640B/50B)

A tten uator 10 dB HP/Agilent 8493C Option 010

(83620B/22B/23B/24B/30B)

Agilent 8360 P erformance T ests

Pulse P erformance (Alternate)

Equipment · Lo cal Oscillator · Mixer P/N 0955-0307

Page 151 of the Agilent Technologies 83640B user manual — ALC
Page 151 · click to enlarge
Page 151
3. On the syn thesizer, press:

CW

First syn thesizer frequency in T able 2-19 .

POWER LEVEL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Leveling Point Internal (asterisk on)

ALC

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Leveling Mode Search (asterisk on)

4. On the lo cal oscillator (83620A/B), set the cen ter frequency to the rst LO frequency in

T able 2-19. Then set:

Power Level10 dBm
RF PowerOn

T able 2-19. Synthesizer and LO Frequencies

Risetime F alltime Lev el

5.0 4.5

9.0 8.5

15.0 14.5

1 Disregard if b eyond your synthesizer's capabilit y .

The CW frequencies c hec k ed provide a minimum v eri cation of pulse

p erformance. If pulse p erformance at a di eren t frequency is imp ortant,

rep eat the test at that CW frequency .

5. On the pulse generator, set:
Pulse Width50 ns
Pulse Period10 s (100 kHz)
Pulse Level5 V
DisableLED off (enables pulse generator)

2-62 P erformance T ests

Pulse P erformance (Alternate)

Show 14 more labels / short text on this page
· · · 5 · · dBm · · · · Synthesizer · LO F requencies · F requencies (GHz) · (GHz) · Accuracy
Page 152 of the Agilent Technologies 83640B user manual — MOD
Page 152 · click to enlarge
Page 152
6. On the oscilloscop e, set:

Channel 1:

Input Impedance 50 ohms

Channel 2:

Timebase:

Delay Reference At center

Trigger:

Trigger Source Chan 2

Trigger Level 2 V

Display:

Number of Averages 1

On the oscilloscop e, turn the w av eform math function on. Then use the

w av eform math maxim um function to determine the pulse en v elop e when

making the follo wing measuremen ts.

7. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Pulse On/Off Extrnl (asterisk on)

MOD

Agilent 8360 P erformance T ests

Pulse P erformance (Alternate)

Show 13 more labels / short text on this page
Display On · Volts/Division 100 mV · Offset 0 V · Input Coupling dc · Display Off · Time/Division 10 ns · Delay 100 ns · Trigger Mode Edge · Display Mode Repetitive · Averaging On · Screen Single · ·
Page 153 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 153 · click to enlarge
Page 153
8. On the oscilloscop e:

a. Adjust the timebase dela y to p osition the rising edge of the pulsed RF near the cen ter

of the display .

b. Adjust the c hannel 1 v olts/division and o set to obtain a 5 division signal lev el

b et w een the RF p ow er o and the RF p ow er on (see Figure 2-19).

c. Adjust the c hannel 1 o set to mov e the RF p ow er o line one-half division b elow the

nearest horizontal graticule (making the horizontal graticule cross the w av eform at the

10% p oint.)

d. Use the oscilloscop e's delta T function to measure the time di erence b et w een the 10%

and 90% risetime p oints on the en v elop e. (The 10% p oint of the risetime is where the

pulse en v elop e crosses the graticule. The 90% p oint is 4 divisions up.) Record this

v alue in T able 2-19.

Agilent Technologies 83640B — 08360125.pdf diagram (page 153)
Fig. 26 · 08360125.pdf

Figure 2-19. Signal Lev el betw een RF P o w er Off and On

9. Adjust the timebase dela y to p osition the falling edge of the pulse near the cen ter of the

oscilloscop e display .

10. Measure the falltime the same w ay as y ou measured the risetime. (See step 8.) Record

this v alue in T able 2-19.

11. Repeat steps 7 through 10 at eac h syn thesizer and LO frequency in T able 2-19 .
12. F or the 83620B /22B and for the 83623B /24B, record eac h w orst case risetime and falltime

v alue from T able 2-19 on the appropriate test record.

Pulse Lev eling Accuracy < 20 GHz

13. On the pulse generator, set:
Pulse Width1 s
Pulse Period1 ms
Pulse Level5 V

2-64 P erformance T ests

Pulse P erformance (Alternate)

Page 154 of the Agilent Technologies 83640B user manual — POWER LEVEL
Page 154 · click to enlarge
Page 154
14. On the syn thesizer, press:

First syn thesizer frequency in T able 2-19

CW

POWER LEVEL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ALC

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Pulse On/Off Extrnl (asterisk on)

MOD

15. On the sp ectrum analyzer, set the cen ter frequency to the rst frequency in T able 2-19.
16. On the oscilloscop e, set:

Channel 1:

Timebase:

Delay Reference At left

Adjust the dela y and timebase so that the pulse w av eform takes up the en tire display .

Adjust the oscilloscop e c hannel 1 o set to p osition the top of the pulse en v elop e near the

cen ter graticule.

17. On the syn thesizer, turn o external pulse and set the p ow er lev el to 0.3 dBm. Press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Pulse On/Off Extrnl (asterisk o )

POWER LEVEL

0.3

Agilent 8360 P erformance T ests

Pulse P erformance (Alternate)

Show 19 more labels / short text on this page
· · · 0 · · · dBm · · · Leveling Mode Normal · · · Volts/Division 30 mV · Time/Division 100 ns · · · · · dBm
Page 155 of the Agilent Technologies 83640B user manual — POWER LEVEL
Page 155 · click to enlarge
Page 155
18. Use the oscilloscop e's w av eform math maximum function to determine the p eak p ow er

lev el (if the w av eform math function w as previously on, clear the display b efore executing

this step.) Use the oscilloscop e's delta V feature and p osition marker 1 at the cen ter of

the p eak p ow er as display ed b y the math w av eform (the width of the w av eform is caused

b y noise in the system.) Lea v e the marker at that p osition.

19. Set the syn thesizer's p ow er lev el to 0.3 dBm. Press:

POWER LEVEL

0.3

20. Clear the oscilloscop e display and use the oscilloscop e's w av eform math maximum feature

to determine the p eak p ow er lev el. Use the oscilloscop e's delta V feature and p osition

marker 2 at the cen ter of the p eak p ow er as display ed b y the math w av eform.

If the di erence b et w een marker 1 and marker 2 is less than 1 division, c hange c hannel 1

v olts/division to a more sensitive scale and rep eat steps 16 through 20.

21. On the syn thesizer, press:

POWER LEVEL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MOD

Pulse On/Off Extrnl (asterisk on)

22. Reset the oscilloscop e w av eform math feature to maxim um and observ e the p eak pulse

p ow er. It should stay within the tw o delta V markers set in steps 17 and 19. Record the

v alue in T able 2-19 .

23. Repeat steps 12 through 21 at eac h syn thesizer and LO frequency in T able 2-19.
24. F or the 83620B /22B and for the 83623B /24B, record the w orst case lev el accuracy from

T able 2-19 on the appropriate test record.

2-66 P erformance T ests

Pulse P erformance (Alternate)

Show 13 more labels / short text on this page
· · · dBm · · · · 0 · · · dBm · ·
Page 156 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 156 · click to enlarge
Page 156

Rise and F all Times > 20 GHz

Disr e gar d if b eyond your synthesizer's c ap ability.

Agilent Technologies 83640B — 08360125.pdf diagram (page 156)
Fig. 27 · 08360125.pdf

Figure 2-20. Pulse P erformance T est Setup >20 GHz

Sp ectrum Analyzer HP/Agilent 8566B

Digitizing Oscilloscop e HP/Agilent 54111D

Micro w av e Ampli er HP/Agilent 11975A

RF Ampli er HP/Agilent 8447F

Pulse Generator HP/Agilent 8116A

Harmonic Mixer HP/Agilent 11970K

Harmonic Mixer HP/Agilent 11970A (83640B/50B)

Harmonic Mixer HP/Agilent 11970Q (83650B)

A tten uator 6 dB HP/Agilent 8493C Option 006

A tten uator 10 dB HP/Agilent 8490D Option 010

Agilent 8360 P erformance T ests

Pulse P erformance (Alternate)

Equipment

Page 157 of the Agilent Technologies 83640B user manual — ALC
Page 157 · click to enlarge
Page 157

T able 2-20. Risetime and F alltime V alues

CW F requency Risetime F alltime Lev el Accuracy

Procedure > 20 to 26.5 GHz.

25. Connect the equipmen t as sho wn in Figure 2-20 using the HP/Agilen t 11970K Mixer.
26. On the syn thesizer, press:

CW

POWER LEVEL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Leveling Point Intrnl (asterisk on)

ALC

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Leveling Mode Search (asterisk on)

27. Set the microw av e ampli er output p ow er lev el to +16 dBm.
28. On the sp ectrum analyzer, select external mixer and set the frequency range for the

mixer. Then set:

Center Frequency23 GHz
Frequency Span0 Hz
29. Repeat steps 5 through 10 recording the risetime and falltime v alues in T able 2-20. F or

83630B only , record the w orst case risetime and w orst case falltime v alues from T able 2-19

and T able 2-20 on the test record.

Procedure > 26.5 to 40 GHz.

30. Replace the HP/Agilent 11970K Mixer with the HP/Agilent 11970A Mixer.
31. Repeat steps 5 through 10 at syn thesizer and sp ectrum analyzer CW frequencies of 30 and

35 GHz, recording the risetime and falltime v alues in T able 2-20. F or 83640B only , record

the w orst case risetime and w orst case falltime v alues from T able 2-19 and T able 2-20 on

the test record.

Procedure > 40 GHz to 50 GHz.

32. Replace the HP/Agilent 11970A Mixer with the HP/Agilent 11970Q Mixer (33 to 50

GHz).

33. Repeat steps 5 through 10 at syn thesizer and sp ectrum analyzer CW frequencies of

45 GHz and record the risetime and falltime v alues in T able 2-20. F or the 83650B

only , record the w orst case risetime and w orst case falltime v alues from T able 2-19 and

T able 2-20 on the test record.

Pulse Lev eling Accuracy > 20 GHz

Disr e gar d if b eyond your synthesizer's c ap ability.

2-68 P erformance T ests

Pulse P erformance (Alternate)

Show 16 more labels / short text on this page
23 GHz · 28 GHz · 35 GHz · 45 GHz · · · 23 · GHz · · · · 0 · dBm · · ·
Page 158 of the Agilent Technologies 83640B user manual — CW
Page 158 · click to enlarge
Page 158

Procedure > 20 to 26.5 GHz.

34. Replace the HP/Agilent 11970Q Mixer with the HP/Agilent 11970K Mixer.
35. On the syn thesizer, press:

CW

36. Repeat steps 13 through 22 and record the lev el accuracy in T able 2-20 .
37. F or 83630B only , record the w orst case lev el accuracy from T able 2-19 and T able 2-20 on

the test record.

Procedure > 26.5 to 40 GHz.

38. Replace the HP/Agilent 11970K Mixer with the HP/Agilent 11970A Mixer.
39. Repeat steps 13 through 22 for CW frequencies of 30 and 35 GHz and record the lev el

accuracy in T able 2-20.

40. F or 83640B only , record the w orst case lev el accuracy from T able 2-19 and T able 2-20 on

the test record.

Procedure > 40 to 50 GHz.

41. Replace the HP/Agilent 11970A Mixer with the HP/Agilent 11970Q Mixer.
42. Repeat steps 13 through 22 for a CW frequency of 45 GHz and record the lev el accuracy

in T able 2-20.

43. F or 83650B only , record the w orst case lev el accuracy from T able 2-19 and T able 2-20 on

the test record.

Agilent 8360 P erformance T ests

Pulse P erformance (Alternate)

Show 5 more labels / short text on this page
· · 23 · · GHz
Page 159 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 159 · click to enlarge
Page 159

In Case of Difficulty

1. A noisy ampli er can cause a noisy trace that is to o wide to measure. Substitute ampli ers

to v erify a problem with the ampli er.

2. If there is no c hange in ampli cation, the ampli er may b e saturated (op erating in

compression). The 0.3 dB v ariation should b e approximately equal on b oth sides. T ry a

low er syn thesizer p ow er lev el to get the ampli er out of saturation.

3. Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

2-70 P erformance T ests

Pulse P erformance (Alternate)

Related Adjustments · None

Page 160 of the Agilent Technologies 83640B user manual — USER CAL
Page 160 · click to enlarge
Page 160
15. Pulse Modulation Video Feedthrough

(F or B-Series Only)

In CW, at sp eci ed maximum lev eled p ow er, the syn thesizer is pulse mo dulated using a

pulse generator. The syn thesizer's RF output is ltered (only the video feedthrough passes),

ampli ed, and display ed on an oscilloscop e. Because of the low amplitude of the video

feedthrough, a preampli er is used in the test setup; system gain m ust b e considered when

making this measuremen t.

When video feedthrough is expressed as a p ercen tage, the follo wing equation is used:

Video feedthrough (%) = [(video Vp/carrier Vp) x 100]/video gain

1. Preset all the instrumen ts sho wn in Figure 2-21 and let them w arm up for at least one

hour.

2. T o ac hiev e p eak p ow er, turn on RF p eaking:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Peak RF Always (asterisk on)

USER CAL

3. Connect the equipmen t as sho wn in Figure 2-21 with the pulse generator connected

directly to c hannel 1 of the oscilloscop e (A).

Agilent Technologies 83640B — USER CAL diagram (page 160)
Fig. 28

Figure 2-21. Video Feedthrough T est Setup

Agilent 8360 P erformance T ests

Pulse Modulation Video Feedthrough

Show 4 more labels / short text on this page
Description and Procedure · · · Tracking Menu
Page 161 of the Agilent Technologies 83640B user manual — (83623B /24B)
Page 161 · click to enlarge
Page 161

Pulse Generator HP/Agilent 8116A

Digitizing Oscilloscop e HP/Agilent 54111D

Preampli er HP/Agilent 8447F

130 MHz Filter K & L Micro w av e

5LL30-130/BT2400/BP

500 MHz Low P ass Filter Mini-Circuits SLP-550

3.7 GHz Filter P/N 9135-0191

A tten uator 6 dB HP/Agilent 8493C Option 006

(83620B /22B/30B/40B/50B)

A tten uator 20 dB HP/Agilent 8493C Option 020

(83623B /24B)

2-72 P erformance T ests

Pulse Modulation Video Feedthrough

Equipment

Page 162 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 162 · click to enlarge
Page 162

The pulse generator output lev el can b e a ected b y loading. Lea v e the pulse

generator connected to the oscilloscop e's c hannel 2 (through the BNC tee) as

y ou set the output lev el.

4. On the oscilloscop e, set:

Channel 1:

Channel 2:

Timebase:

Delay Reference At center

Trigger:

Trigger Source Channel 1

Trigger Level 0.05 V

Display:

Number of Averages 16

5. On the pulse generator, set:
PulseSelected
Pulse width10 ns
Frequency2 MHz (500 ns)
Offset0 V
6. On the pulse generator, set the pulse amplitude for a reading of 0.1 V on the oscilloscop e.
7. Connect the pulse generator to the oscilloscop e's c hannel 1 through the attenuator, lters,

and ampli er (B).

8. On the oscilloscop e, set:

Channel 1:

Volts/Division 0.2 V

Offset 0.3 V

Agilent 8360 P erformance T ests

Pulse Modulation Video Feedthrough

Show 16 more labels / short text on this page
System Calibration · Display On · Volts/Division 50 mV · Offset 0 V · Input Coupling dc · Input Impedance 50 · Display Off · Input Impedance 50 · Time/Division 20 ns · Delay 0 s · Trigger Mode Edge · Trigger Slope Positive · Display Mode Repetitive · Averaging On · Video Gain · Input Coupling dc
Page 163 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 163 · click to enlarge
Page 163

Trigger:

Trigger Source Channel 1

Trigger Level 0.4 V

9. On the oscilloscop e, note the pulse amplitude: Vp
10. Calculate the video gain:

system Vp (from step 9)

0. 1 Vp

Lo w Band Video Feedthrough

Disr e gar d if b eyond your synthesizer's c ap ability.

11. Connect the pulse generator output to the syn thesizer's pulse input (C).
12. Connect the syn thesizer to the oscilloscop e's c hannel 1 through the attenuator, lters, and

ampli er (D).

13. On the syn thesizer, press:

First CW frequency in T able 2-21

CW

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ALC

Leveling Mode Search .

POWER LEVEL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MOD

Pulse On/Off Extrnl (asterisk on)

or, for syn thesizers with Option 002, instead press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MOD

T able 2-21. Lo w Band Video Feedthrough Frequencies

2-74 P erformance T ests

Pulse Modulation Video Feedthrough

Show 31 more labels / short text on this page
Input Impedance 50 · Trigger Mode Edge · Trigger Slope Negative · Video gain = · pulse generator Vp · system Vp · = · Video Gain = · · · · · · 10 · · dBm · · · · · Pulse Menu · Pulse On/Off Ext · Syn thesizer CW · Video · Video · F requency · F eedthrough (%) · F eedthrough (%) · (GHz) · 10 dBm · > 10 dBm
Page 164 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 164 · click to enlarge
Page 164
14. On the oscilloscop e, set:

Channel 1:

Volts/Division 0.01 V

Channel 2:

Timebase:

Delay Reference At center

Trigger:

Trigger Source Channel 2

Trigger Level 1 V

Display:

Number of averages 64

15. On the pulse generator, set:
Pulse width500 ns
Frequency200 kHz (5 s)
Offset0 V
Amplitude5 V
16. On the oscilloscop e, note the maximum v oltage from the baseline.

(See Figure 2-22.) Video feedthrough (Vp) = Vp

Agilent 8360 P erformance T ests

Pulse Modulation Video Feedthrough

Show 13 more labels / short text on this page
Display On · Offset 0 V · Input Coupling dc · Input Impedance 50 · Display Off · Volts/Division 1 V · Offset 2 V · Time/Division 200 ns · Delay 400 ns · Trigger Mode Edge · Trigger Slope Positive · Display Mode Repetitive · Averaging On
Page 165 of the Agilent Technologies 83640B user manual — (V)
Page 165 · click to enlarge
Page 165
Agilent Technologies 83640B — 08360125.pdf diagram (page 165)
Fig. 29 · 08360125.pdf

Figure 2-22. Measuring Maximum V oltage from the Baseline Video Feedthrough

17. Using the video gain from step 10, the video feedthrough (Vp) from step 16, and the

carrier v oltage from T able 2-22 calculate the video feedthrough as a p ercent of carrier

p ow er:

Video feedthrough (%) = [(video Vp/carrier Vp)x100]/video gain

T able 2-22. P o w er Lev el Conv ersions to V olts P eak

Sp eci ed Po w er

Sp eci ed Po w er

(V)

(V)

0 0.316

10.5 1.059

0.5 0.335

11.0 1.122

1.0 0.355

11.5 1.189

1.5 0.376

12.0 1.259

2.0 0.398

12.5 1.334

2.5 0.422

13.0 1.413

3.0 0.447

13.5 1.496

3.5 0.473

14.0 1.585

4.0 0.501

14.5 1.679

4.5 0.531

15.0 1.778

5.0 0.562

15.5 1.884

5.5 0.596

16.0 1.995

6.0 0.631

16.5 2.113

6.5 0.668

17.0 2.239

7.0 0.708

17.5 2.371

7.5 0.750

18.0 2.512

8.0 0.794

18.5 2.661

8.5 0.841

19.0 2.819

9.0 0.891

19.5 2.985

9.5 0.944

20.0 2.985

10.0 1.000

2-76 P erformance T ests

Pulse Modulation Video Feedthrough

Show 12 more labels / short text on this page
Synthesizer Maxim um · Peak · Synthesizer Maxim um · Peak · Carrier · Carrier · (dBm) · V oltage · (dBm) · V oltage · In to 50 · In to 50
Page 166 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 166 · click to enlarge
Page 166

P eak Carrier V oltage = 10

18. Record the video feedthrough (in T able 2-21).
19. Repeat steps 16 through 18 for the remaining CW frequencies in T able 2-21 .
20. F or syn thesizers capable of > 10 dBm, set the maxim um sp eci ed lev eled p ow er and

rep eat steps 16 through 18 for all of the CW frequencies in T able 2-21 .

21. Record the w orst case v alue from T able 2-21 for eac h p ow er range on the test record.

Video Feedthrough > 2 GHz

22. Replace the 130 MHz low pass lter with a 500 MHz low pass lter.
23. Repeat steps 7 through 10 to c hec k video gain once more.
24. On the oscilloscop e, reset c hannel 1 to 0.01 v olts/division.
25. Set the syn thesizer to its maximum sp eci ed lev eled p ow er and rep eat steps 13 and

16 through 17 (the oscilloscop e and pulse generator are already set up) for the CW

frequencies in T able 2-23 . Do not exceed the maxim um frequency of y our syn thesizer.

Record the video feedthrough in T able 2-23 .

T able 2-23. Millimeter Band Video Feedthrough Frequencies

26. Record the w orst case v alue in T able 2-23 on the test record.

In Case of Difficulty

Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series Swept

CW Gener ator T r oublesho oting Guide .

Agilent 8360 P erformance T ests

Pulse Modulation Video Feedthrough

Show 9 more labels / short text on this page
· · PdBm10 · F requency · Video · (GHz) · F eedthrough (%) · Related Adjustments · None
Page 167 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 167 · click to enlarge
Page 167
16. AM Accuracy

(F or B-Series Only)

This pro cedure measures the AM accuracy of the syn thesizer. The lo cal oscillator and mixer

are used to do wncon v ert the syn thesizer frequency to the frequency range of the measuring

receiv er (an HP/Agilent 11793A Micro w av e Con v erter can b e substituted for the mixer).

F or measuremen ts ab ov e 20 GHz RF output frequency , a sp ectrum analyzer with external

mixer are used to do wncon v ert to a 321.4 MHz RF. The function generator provides AM

mo dulation. The function generator signal lev el is adjusted for 30% AM as measured b y the

measuring receiv er. The function generator signal lev el is c hec k ed for accuracy . A DVM may

b e necessary to measure the function generator output.

1. T urn on the equipmen t sho wn in Figure 2-23 . Preset the instrumen ts and let them w arm

up for at least one hour.

2. On the measuring receiv er, calibrate and store the AM calibration factor.

Lo w Band < 2 GHz

Disr e gar d if b eyond your synthesizer's c ap ability.

3. Connect the equipmen t as sho wn in Figure 2-23.
Agilent Technologies 83640B — 08360125.pdf diagram (page 167)
Fig. 30 · 08360125.pdf

Figure 2-23. AM Accuracy T est Setup Lo w Band

2-78 P erformance T ests

Description and Procedure · AM Accuracy

Page 168 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 168 · click to enlarge
Page 168

F unction Generator HP/Agilent 3325A

Digital V oltmeter HP/Agilent 3456A

Measuring Receiv er HP/Agilent 8902A

A tten uator 10 dB HP/Agielnt 8490D Option 010 (83640B/50B)

A tten uator 10 dB HP/Agilent 8493C Option 010 (83620B/23B/30B)

4. On the syn thesizer, press:

CW

POWER LEVEL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM BW Cal Always (asterisk on)

USER CAL

5. On the function generator, set:
FunctionSine wave
Frequency1 kHz
DC Offset0 mV

Set the amplitude to 600 mV

) and v erify this amplitude with the DVM.

(V

x 1.414 x 2 = V

).

6. On the measuring receiv er, set:
HP Filter300 Hz
LP Filter3 kHz
MeasurementAM
DetectorPeak /2 (both on)
7. On the syn thesizer, turn on AM mo dulation. Press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MOD

AM On/Off 100%/V (asterisk on)

F or instrumen ts with Option 002, instead press:

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM On/Off Ext (asterisk on)

MOD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM Type 100%/V (asterisk on)

8. On the test record, record the mo dulation depth as indicated b y the measuring receiv er.

For Instruments with Option 002 only .

9. On the syn thesizer, turn on in ternal AM. Press:

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM On/Off Int (asterisk on)

MOD

10. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Internal AM Rate

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Internal AM Depth

11. On the test record, record the mo dulation depth as indicated b y the measuring receiv er.

Agilent 8360 P erformance T ests

Show 35 more labels / short text on this page
Equipment · · · 500 · MHz · · · · 0 · dBm · · · · AM Cal Menu · p-p · (or 212 mV · rms · rms · p-p · Automatic Operation · · · · · AM Menu · · · AM Menu · 1 · kHz · · 30 · enter · · AM Accuracy
Page 169 of the Agilent Technologies 83640B user manual — CW
Page 169 · click to enlarge
Page 169

High Band > 2 GHz and < 20 GHz

12. Connect the equipmen t as sho wn in Figure 2-24.
Agilent Technologies 83640B — 08360125.pdf diagram (page 169)
Fig. 31 · 08360125.pdf

Figure 2-24. AM Accuracy T est Setup < 20 GHz

F unction Generator HP/Agilent 3325A

Digital V oltmeter HP/Agilent 3456A

(Syn thesized Sw eeper) 83620A/B

Measuring Receiv er HP/Agilent 8902A

A tten uator 10 dB HP/Agilent 8490D Option 010

(83640B/50B)

A tten uator 10 dB HP/Agilent 8493C Option 010

(83620B/22B/23B/24B/30B)

13. On the syn thesizer, press:

CW

POWER LEVEL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USER CAL

AM BW Cal Always (asterisk on)

14. On the lo cal oscillator, set:
CW Frequency5.1 GHz
RFOn
Power Level10 dBm

2-80 P erformance T ests

Show 17 more labels / short text on this page
Equipment · Lo cal Oscillator · Mixer P/N 0955-0307 · · · 5 · GHz · · · · 0 · · dBm · · · AM Cal Menu · AM Accuracy
Page 170 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 170 · click to enlarge
Page 170
15. On the function generator, set:
FunctionSine wave
Frequency1 kHz
DC Offset0 mV

Set the amplitude to 600 mV p-p and v erify this amplitude with the DVM. Note that the

DVM measures in V rms (V rms x 1.414 x 2 = V p-p).

16. On the measuring receiv er, set:
HP Filter300 Hz
LP Filter3 kHz
MeasurementAM
DetectorPeak /2 (both on)
17. On the syn thesizer, turn on AM mo dulation. Press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MOD

AM On/Off 100%/V (asterisk on)

F or instrumen ts with Option 002, instead press:

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM On/Off Ext (asterisk on)

MOD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM Type 100%/V (asterisk on)

18. On the test record, record the mo dulation depth as indicated b y the measuring receiv er.

For Instruments with Option 002 only .

19. On the syn thesizer, turn external AM o and in ternal AM on. Press:

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM On/Off Ext (asterisk o )

MOD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM On/Off Int (asterisk on)

20. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Internal AM Rate

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Internal AM Depth

21. On the test record, record the mo dulation depth as indicated b y the measuring receiv er.

High Band > 20 GHz { Doubler Engaged

Disr e gar d if b eyond your synthesizer's c ap ability.

22. Connect the equipmen t as sho wn in Figure 2-25.

Agilent 8360 P erformance T ests

Show 16 more labels / short text on this page
Automatic Operation · · · · · AM Menu · · · AM Menu · 1 · kHz · · 30 · · enter · AM Accuracy
Page 171 of the Agilent Technologies 83640B user manual — MOD
Page 171 · click to enlarge
Page 171
Agilent Technologies 83640B — 08360125.pdf diagram (page 171)
Fig. 32 · 08360125.pdf

Figure 2-25. AM Accuracy T est Setup > 20 GHz

Sp ectrum Analyzer HP/Agilent 8566B

Micro w av e Ampli er HP/Agilent 11975A

Measuring Receiv er HP/Agilent 8902A

Harmonic Mixer P/N 0955-0307

F unction Generator HP/Agilent 3325A

A tten uator 10 dB HP/Agilent 8490D Option 010

23. On the syn thesizer, press:

CW

POWER LEVEL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MOD

AM On/Off 100%/V (asterisk o )

F or instrumen ts with Option 002, instead press:

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MOD

AM On/Off Ext (asterisk on)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM Type 100%/V (asterisk on)

24. Set the microw av e ampli er output p ow er lev el to +16 dBm.
25. On the sp ectrum analyzer, set:
Center Frequency28 GHz
Frequency Span0 Hz
26. Set up the function generator as in step 6.
27. Set up the measuring receiv er as in step 7 and select automatic op eration.
28. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM On/Off 100%/V (asterisk on)

MOD

2-82 P erformance T ests

Show 20 more labels / short text on this page
Equipment · · · 28 · · GHz · · · · · 3 · dBm · · · · · AM Menu · · · AM Accuracy
Page 172 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 172 · click to enlarge
Page 172

F or instrumen ts with Option 002, instead press:

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM On/Off Ext (asterisk on)

MOD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM Type 100%/V (asterisk on)

29. On the test record, record the mo dulation depth as indicated b y the measuring receiv er.

For Instruments with Option 002 only:.

30. On the syn thesizer, turn external AM o and in ternal AM on. Press:

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM On/Off Ext (asterisk o )

MOD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM On/Off Int (asterisk on)

31. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Internal AM Rate

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Internal AM Depth

32. On the test record, record the mo dulation depth as indicated b y the measuring receiv er.

Mo dulator O set and Gain

In Case of Difficulty

1. The AM input imp edance is in ternally selectable to either 50 or 2 k

(the factory-set v alue is 50 ). If the AM input is set for 2 k , and the function generator

requires a 50 system, use a DVM to measure the AM input signal lev el while it is

connected to the syn thesizer.

2. Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

Agilent 8360 P erformance T ests

Show 14 more labels / short text on this page
· · AM Menu · · · AM Menu · 1 · kHz · · 30 · enter · · Related Adjustments · AM Accuracy
Page 173 of the Agilent Technologies 83640B user manual — PRESET
Page 173 · click to enlarge
Page 173
17. AM Bandwidth

(F or B-Series Only)

This test v eri es that the p eak RF output p ow er do es not v ary more than a sp eci ed amoun t

ov er the sp eci ed amplitude mo dulation rate frequency range. The sp ectrum analyzer with

the tracking generator op erates as a net w ork analyzer to measure the atness. Flatness

errors asso ciated with the tracking generator and sp ectrum analyzer are calibrated out of the

measuremen t.

1. Connect the equipmen t as sho wn in Figure 2-26 with the tracking generator output

connected to the 50 ohm input of the sp ectrum analyzer and to the DVM. Press

PRESET

and let the equipmen t w arm up for at least one hour.

Agilent Technologies 83640B — PRESET diagram (page 173)
Fig. 33

Figure 2-26. AM Bandwidth T est Setup

With T rac king Generator HP/Agilent 3585A/B

Crystal Detector HP/Agilent 33330E (83640B/50B)

Crystal Detector HP/Agilent 33330D (83620B/22B/23B/24B/30B)

Digital V oltmeter HP/Agilent 3456A

2. On the sp ectrum analyzer, set:
Start Frequency5 kHz
Stop Frequency200 kHz
Sweep Manual5 kHz

2-84 P erformance T ests

Show 5 more labels / short text on this page
Description and Procedure · · Equipment · Sp ectrum Analyzer · AM Bandwidth
Page 174 of the Agilent Technologies 83640B user manual — MOD
Page 174 · click to enlarge
Page 174
dB/Div1 dB
Range5 dBm
AmplitudeMaximum 600 mV p-p (212 mV rms on the DVM)
3. On the sp ectrum analyzer, clear trace A memory . Change the

REF LEV

, using the RPG

knob, to p osition the trace (dot) on the cen ter horizontal graticule.

4. Set the sp ectrum analyzer to con tinuous sweep. When a full sweep is completed, store

trace A in to trace B memory . Then display A minus B (turning o trace B). A straight

line should b e display ed.

The sp ectrum analyzer is no w calibrated.

5. Connect the tracking generator output to the syn thesizer's AM input and connect the

syn thesizer's RF output through the crystal detector to the sp ectrum analyzer's 50 ohm

input.

6. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM BW CAL Always (asterisk on)

USER CAL

AM CAL MENU

CW

First syn thesizer frequency in T able 2-24

POWER LEVEL

Set to 5 dB b elow sp eci ed maximum lev eled p ow er

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM On/Off 100%/V (asterisk on)

MOD

or for syn thesizers with Option 002, press:

MOD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

T able 2-24. Synthesizer Frequencies

1 Disregard if b ey ond your synthesizer's

capability .

7. On the sp ectrum analyzer, clear and view trace A (ligh t on). Change the reference lev el to

set the p eak of the trace on the cen ter graticule. If an ov er range o ccurs, turn the RPG

knob in the opp osite direction.

The display no w sho ws the syn thesizer AM atness from 5 kHz to 200 kHz.

8. Use the sp ectrum analyzer markers to determine the 3 dB bandwidth. Record the

frequency of the 3 dB bandwidth in T able 2-24. If the 3 dB bandwidth p oint is o the

display , record \ > 200 kHz" in T able 2-24.

Agilent 8360 P erformance T ests

Show 16 more labels / short text on this page
Tracking Generator · · · · · · · · · AM On/Off Ext · AM Type 100%/V · Syn thesizer · 3 dB · F requencies (GHz) · Bandwidth · AM Bandwidth
Page 175 of the Agilent Technologies 83640B user manual — CONT
Page 175 · click to enlarge
Page 175
9. Set the sp ectrum analyzer to

and rep eat steps 7 and 8 for the remaining

CONT

syn thesizer frequencies in T able 2-24.

10. Record the smallest bandwidth from T able 2-24 on the test record.

In Case of Difficulty

Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

2-86 P erformance T ests

Related Adjustments · None · AM Bandwidth

Page 176 of the Agilent Technologies 83640B user manual — CA UTION
Page 176 · click to enlarge
Page 176
18. AM Dynamic Range

(F or B-Series Only)

In this pro cedure, a function generator is used to provide a DC v oltage to the syn thesizer's

AM input. The syn thesizer is set to 0 dBm output p ow er. AM is enabled and a DC v oltage

is applied. The DC v oltage is set for maximum AM without an ov ermo dulation indication to

drive the syn thesizer's mo dulator to the edge of distortion. The output p ow er is measured

with a p ow er meter.

A measuremen t is made in 1 GHz steps across the syn thesizer's frequency range. This

equipmen t lets y ou measure do wn to 30 dBm. If y ou reac h this lev el b efore the syn thesizer

ov ermo dulates, the p ow er meter will indicate an under range measuremen t.

This pro cedure is rep eated using the deep AM function in the searc h ALC mo de. A sp ectrum

analyzer is used in place of the p ow er meter to make the deep AM measuremen t. This

equipmen t allo ws y ou to measure do wn to approximately 65 dBm. If y our syn thesizer

do es not ov ermo dulate b y this p oint, and y ou wish to measure the p ow er lev el just b efore

ov ermo dulation, use a p ow er supply to increase the DC v oltage supply to the AM input.

CA UTION

Do not exceed 15 V DC to the AM input or damage will o ccur.

1. Preset the instrumen ts sho wn in Figure 2-27 and let them w arm up for at least one hour.
Agilent Technologies 83640B — CA UTION diagram (page 176)
Fig. 34

Figure 2-27. AM Dynamic Range T est Setup

Agilent 8360 P erformance T ests

Description and Procedure · AM Dynamic Range

Page 177 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 177 · click to enlarge
Page 177

Sp ectrum Analyzer HP/Agilent 8566B

Digital V oltmeter HP/Agilent 3456A

F unction Generator HP/Agilent 3325A

P ow er Meter HP/Agilent 436A/7A/8A

P ow er Sensor HP/Agilent 8487A (83640B/50B)

P ow er Sensor HP/Agilent 8485A (83620B/22B/23B/24B/30B)

2. On the syn thesizer, turn on RF p eaking and AM bandwidth calibration. Press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USER CAL

Peak RF Always (asterisk on)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRIOR

AM BW Cal Always (asterisk on)

3. Zero and calibrate the p ow er meter/sensor.
4. Connect the equipmen t as sho wn in Figure 2-27 with the p ow er sensor connected to the

syn thesizer's RF OUTPUT.

5. On the syn thesizer, press:

CW

(Set to 2 GHz for syn thesizers with low est start frequency of 2 GHz)

FREQUENCY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Up/Dn Size CW

MENU

PO WER

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Uncoupl Atten (Option 001 only)

POWER LEVEL

MOD

V erify that all mo dulation is o . (No asterisks next to k ey lab els.)

F or syn thesizers with Option 002, if mo dulation is on, an \AM," \FM," or \PLS" message

will b e display ed.

6. On the p ow er meter, set:
ModedBm
Cal Factor %Cal factor corresponding to the
7. On the function generator, turn o all w av eforms and set:
DC Offset 1.5 V DC

Normal AM Dynamic Range

8. On the syn thesizer, turn on AM mo dulation to 10 dB/V:

MOD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM On/Off 10dB/V (asterisk on)

or, for syn thesizers with Option 002, press:

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MOD

AM On/Off Ext (asterisk on)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM Type 10 dB/V (asterisk on)

9. Decrease the function generator output just un til the O VERMOD message turns on.

Then increase the signal un til the O VERMOD message turns o . (The more negative the

signal applied to the mo dulator is, the more the RF output level decreases.)

10. W ait for the p ow er meter to settle then note the p ow er meter reading in T able 2-25. If the

p ow er meter reads under range, record < 30 dB.

2-88 P erformance T ests

Show 30 more labels / short text on this page
Equipment · · · Tracking Menu · · · AM Cal Menu · · · 1 · GHz · · 1 · · GHz · · · · 10 · · · dBm · · synthesizer's CW frequency · · · · · AM Menu · AM Dynamic Range
Page 178 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNN
Page 178 · click to enlarge
Page 178

T able 2-25. P o w er Meter Readings

1 GHz 12 GHz

2 GHz 13 GHz

3 GHz 14 GHz

4 GHz 15 GHz

5 GHz 16 GHz

6 GHz 17 GHz

7 GHz 18 GHz

8 GHz 19 GHz

9 GHz 20 GHz

10 GHz 21 GHz

11 GHz 22 GHz

1 83630B/40B/50B only .

11. Repeat steps 9 and 10 from 2 to 22 GHz, in 1 GHz steps. Change the p ow er meter cal

factor % as required.

12. Record the w orst case v alue from T able 2-25 on the test record.

Deep AM Dynamic Range

13. Disconnect the p ow er sensor from the syn thesizer and connect the sp ectrum analyzer as

sho wn in Figure 2-27.

14. On the syn thesizer, press:

CW

(Set to 2 GHz for syn thesizers with low est start frequency of 2 GHz)

POWER LEVEL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM On/Off 10 dB/V (asterisk o )

or for syn thesizers with Option 002, press:

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM On/Off Int (asterisk o )

MOD

15. On the sp ectrum analyzer, set:
Center Frequency1 GHz
Frequency Span1 MHz
CF Step Size1 GHz
Reference Level5 dBm
Scale Log5 dB/Division
16. Cen ter the signal on the sp ectrum analyzer with the cen ter frequency con trol.
17. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM On/Off 100%/V (asterisk on)

NNNNNNNNNNNNNNNNNNNNNNN

Deep AM (asterisk on)

or for syn thesizers with Option 002, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM Type 100%/V (asterisk o )

NNNNNNNNNNNNNNNNNNNNNNN

Deep AM (asterisk on)

18. Change the sp ectrum analyzer reference lev el to 40 dBm.

Agilent 8360 P erformance T ests

Show 15 more labels / short text on this page
· · 1 · GHz · · · 15 · · · dBm · · · AM Menu · AM On/Off Ext · AM Dynamic Range
Page 179 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 179 · click to enlarge
Page 179

Make the follo wing measuremen ts as quic kly as p ossible since the signal

amplitude is sub ject to drift when y ou use deep AM at low lev els. If necessary ,

turning AM o and on again will reset the p ow er lev el.

19. With a 10 mV resolution, adjust the function generator un til y ou get an O VERMOD

message on the syn thesizer or un til the signal drops in to the noise lev el on the sp ectrum

analyzer. Decrease the function generator output just un til the O VERMOD message

disapp ears.

20. Record the RF signal lev el (or the noise lev el if the signal is in the noise).

T able 2-26. RF Signal Lev el

1 GHz 12 GHz

2 GHz 13 GHz

3 GHz 14 GHz

4 GHz 15 GHz

5 GHz 16 GHz

6 GHz 17 GHz

7 GHz 18 GHz

8 GHz 19 GHz

9 GHz 20 GHz

10 GHz 21 GHz

11 GHz 22 GHz

1 83630B/40B/50B only .

21. On the syn thesizer, turn o mo dulation. Set:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM On/Off 10dB/V (asterisk o )

or for syn thesizers with Option 002, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM Type 10 dB/V (asterisk o )

22. Step the syn thesizer and sp ectrum analyzer CW and cen ter frequencies in 1 GHz steps.

T urn mo dulation bac k on and rep eat steps 19 through 21 to the highest frequency from

T able 2-26 that applies to y our syn thesizer.

23. Record the w orst case v alues from step 20 on the test record:

a. F or the 83630B , record w orst case v alues for frequencies < 20 GHz and for frequencies

20 GHz.

b. F or the 83640B /50B, record w orst case v alues for frequencies 20 GHz and for

frequencies > 20 GHz.

c. F or all other instrumen ts, record the w orst case v alue for all frequencies tested.

2-90 P erformance T ests

AM Dynamic Range

Page 180 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 180 · click to enlarge
Page 180

Mo dulator O set and Gain

Mo dulation Generator Adjustmen t (Option 002)

In Case of Difficulty

1. Make sure that the most negative v oltage without an O VERMOD message is applied.
2. Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

Agilent 8360 P erformance T ests

Related Adjustments · AM Dynamic Range

Page 181 of the Agilent Technologies 83640B user manual — POWER LEVEL
Page 181 · click to enlarge
Page 181
19. FM Accuracy

(F or B-Series Only)

Use this pro cedure to measure the FM accuracy of the syn thesizer. With the syn thesizer set

to an FM sensitivit y of 10 MHz/V, and the function generator providing an FM mo dulation

index of 2.404, the RF output should b e a n ull. The function generator amplitude is adjusted

for the Bessel n ull and the di erence b et w een signals is recorded.

Since the mo dulation frequency for external FM is 1 MHz, this pro cedure

relies on the v oltage-setting accuracy of the function generator (the frequency

is to o high for a DVM, and an oscilloscop e is not accurate enough.)

1. Connect the equipmen t as sho wn in Figure 2-28. Preset all instrumen ts and let them

w arm up for at least one hour.

Agilent Technologies 83640B — 08360125.pdf diagram (page 181)
Fig. 35 · 08360125.pdf

Figure 2-28. FM Accuracy T est Setup

Sp ectrum Analyzer HP/Agilent 8566B

F unction Generator HP/Agilent 3325A

2. T o ac hiev e p eak p ow er on the syn thesizer, turn on RF p eaking. Press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USER CAL

Peak RF Always (asterisk on)

3. On the syn thesizer, press:

First syn thesizer frequency in T able 2-27

CW

Set the sp eci ed maximum lev eled p ow er

POWER LEVEL

2-92 P erformance T ests

Show 9 more labels / short text on this page
Description and Procedure · External FM · Equipment · · · Tracking Menu · · · FM Accuracy
Page 182 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 182 · click to enlarge
Page 182

T able 2-27. Synthesizer Frequencies

Syn thesizer F requencies

1 Disregard if b eyond your synthesizer's capability .

4. On the function generator, set:
FunctionSine wave
Frequency1 MHz
Amplitude480.8 mV p-p (169.96 mV rms)
DC Offset0 V
5. On the sp ectrum analyzer, set:
Reference Level0 dBm
AttenuatorAuto
Center FrequencySame frequency as in step 3
Frequency Span50 kHz
Resolution Bandwidth3 kHz
Video Bandwidth1 kHz
Sweep TimeAuto
SweepContinuous
6. Cen ter the signal on the sp ectrum analyzer display using the cen ter frequency function.
7. T urn on the syn thesizer mo dulation. Press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MOD

FM On/Off 100 kHz (asterisk on)

or for syn thesizers with Option 002, press:

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FM On/Off Ext (asterisk on)

MOD

8. Adjust the function generator amplitude to minimize the amplitude of the signal on the

sp ectrum analyzer.

9. Calculate the p ercen tage of error as follows:

Error (%) = [(480.8 mV p-p New Amplitude)/480.8 mV p-p] x 100.

10. Record the error in T able 2-27 .
11. T urn o the syn thesizer FM mo dulation. Press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FM On/Off 100 kHz (asterisk o )

or for syn thesizers with Option 002, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FM On/Off Ext (asterisk o )

12. Repeat steps 6 through 11 at the syn thesizer (and sp ectrum analyzer) frequencies in

T able 2-27.

13. Record the w orst case v alue from T able 2-27 on the test record.

Agilent 8360 P erformance T ests

Show 11 more labels / short text on this page
External FM · In ternal FM · (GHz) · Error (%) · Error (%) · · · · · FM Menu · FM Accuracy
Page 183 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 183 · click to enlarge
Page 183

Internal FM (Option 002 only)

14. On the syn thesizer, press:

First syn thesizer frequency in T able 2-27

CW

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MOD

FM On/Off Ext (asterisk o )

15. Set the sp ectrum analyzer cen ter frequency to the same frequency as the syn thesizer.
16. Cen ter the signal on the sp ectrum analyzer display using the cen ter frequency function.
17. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MOD

FM On/Off Int (asterisk on)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Internal FM Rate

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Internal FM Dev

2.404

18. Adjust the in ternal FM deviation on the syn thesizer to minimize the amplitude of the

carrier on the sp ectrum analyzer.

19. Calculate the p ercen tage of error as follows:

Error (%) = [(2.404 MHz New FM Deviation)/2.404 MHz] x 100.

20. Record the error in T able 2-27.
21. T urn o the in ternal FM mo dulation. Select:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FM On/Off Int (asterisk o )

22. Repeat steps 15 through 21 at the syn thesizer (and sp ectrum analyzer) frequencies in

T able 2-27.

23. Record the w orst case \In ternal FM Error" (from T able 2-27 ) on the test record.

Mo dulation Generator (Option 002)

Mo dulation Generator Flatness (Option 002)

In case of Difficulty

Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

2-94 P erformance T ests

Show 15 more labels / short text on this page
· · · FM Menu · · · FM Menu · 1 · MHz · · · MHz · Related Adjustments · FM Gain · FM Accuracy
Page 184 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 184 · click to enlarge
Page 184
20. FM Bandwidth

(F or B-Series Only)

Use this pro cedure to v erify that the RF output p ow er do es not v ary more than a sp eci ed

amoun t ov er the sp eci ed frequency mo dulation rate frequency range. A low frequency

sp ectrum analyzer with a tracking generator op erates as a net w ork analyzer to measure the

FM atness. Flatness errors asso ciated with the tracking generator and sp ectrum analyzer are

calibrated out of the measuremen t. Flatness errors due to the p ow er splitter and microw a v e

ampli er are minimized b y external lev eling at the ampli er. The dela y line discriminator is

used to demo dulate the FM from the RF carrier. It also provides an input to the sp ectrum

analyzer that is equiv alen t to the tracking generator frequency and prop ortional in p ow er to

the syn thesizer RF output.

The dela y line discriminator is formed b y driving a mixer with a mo dulated carrier in to

the LO p ort, and a dela y ed carrier in to the RF p ort. The dela y is pro duced with a cable of

approximately 3 feet in length. With this setup, the mixer resp onds to phase di erences

b et w een the tw o inputs. The cable has a constant time dela y and, as the carrier frequency is

frequency-mo dulated, a linear phase shift o ccurs b et w een the mixer inputs. Since the mixer

inputs are in quadrature, the mixer op erates as a phase detector that amplitude tracks the

RF input (syn thesizer p ow er) and has an output frequency corresp onding to the FM rate

(tracking generator frequency).

Agilent 8360 P erformance T ests

Description and Procedure · FM Bandwidth

Page 185 of the Agilent Technologies 83640B user manual — USER CAL
Page 185 · click to enlarge
Page 185
1. Connect the equipmen t as sho wn in Figure 2-29 with the tracking generator output

connected to the c hannel 1 input of the oscilloscop e and to the FM input of the

syn thesizer (A). A DVM can b e used in place of the oscilloscop e for rms measuremen ts.

Preset all the equipmen t and let them w arm up at least one hour.

2. T o ac hiev e p eak p ow er on the syn thesizer, turn on RF p eaking. Press:

USER CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Peak RF Always (asterisk on)

3. On the sp ectrum analyzer, set:
Center frequency100 kHz
Frequency Span0 Hz

Set the sp ectrum analyzer tracking generator to 100 mV p-p (35 mV rms). Cho ose the

alternate detector if y our tracking generator has an alternate detector that improv es the

low-end atness.

4. Connect the tracking generator output to the 50 ohm input of the sp ectrum analyzer

through the 10 dB attenuator (B). On the sp ectrum analyzer, set:

Start Frequency100 kHz
Stop Frequency10 MHz
Resolution Bandwidth10 kHz
Video Bandwidth1 kHz
SweepContinuous
dB/Div1 dB
Range 10 dBm

2-96 P erformance T ests

Show 4 more labels / short text on this page
· · Tracking Menu · FM Bandwidth
Page 186 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 186 · click to enlarge
Page 186
Agilent Technologies 83640B — 08360125.pdf diagram (page 186)
Fig. 36 · 08360125.pdf

Figure 2-29. FM Bandwidth T est Setup

Agilent 8360 P erformance T ests

FM Bandwidth

Page 187 of the Agilent Technologies 83640B user manual — CW
Page 187 · click to enlarge
Page 187

Sp ectrum Analyzer with

T rac king Generator

Digitizing Oscilloscop e HP/Agilent 54111D

Digital V oltmeter HP/Agilent 3456A

Micro w av e Ampli er HP/Agilent 8348A

(83620B/22B/23B/24B/30B)

P ow er Splitter HP/Agilent 11667C (83640B/50B)

P ow er Splitter HP/Agilent 11667B

(83620B/22B/23B/24B/30B)

A tten uator 10 dB (2) HP/Agilent 8490D Option 010

(83640B/50B)

A tten uator 10 dB (2) HP/Agilent 8493C Option 010

(83620B/22B/23B/24B/30B)

P/N 08503-20038

Discriminator (cable 3 ft.)

5. On the sp ectrum analyzer, clear trace A memory . Change the reference lev el, using the

RPG knob, to p osition the trace on the cen ter horizontal graticule.

6. When a full sweep is completed, store trace A in to trace B memory . Then display A

minus B (turning o trace B). A straight line should b e display ed.

7. Connect the tracking generator output to the syn thesizer's FM input and connect the

mixer's IF output to the DVM's V OL TS HI/LO input (C).

Set the DVM to measure DC v olts.

8. On the syn thesizer, press:

First syn thesizer frequency in T able 2-28

CW

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ALC

POWER LEVEL

Set the p ow er so that the microw av e ampli er displays an output of 16

dBm.

T able 2-28. Synthesizer Frequencies

Syn thesizer F requencies

1 Disregard if b eyond your synthesizer's capability .

2 83630B/40B/50B only .

2-98 P erformance T ests

Show 18 more labels / short text on this page
Equipment · HP/Agilent 3585A/B · Mixer P/N 0955-0307 · Delay Line · Find Quadrature · · · · Leveling Point ExtDet · Coupling Factor · 16 · dB · · · Maximum · (GHz) · Di erence · FM Bandwidth
Page 188 of the Agilent Technologies 83640B user manual — MOD
Page 188 · click to enlarge
Page 188
9. On the sp ectrum analyzer, set:
Center Frequency100 kHz
Frequency Span0 Hz
10. On the syn thesizer, press

CW

and use the left arrow k ey to p osition the cursor in the 1

MHz p osition (one digit left of the decimal p oint). Use the rotary knob or step k eys to

adjust the CW frequency for 0 V DC on the DVM.

DC FM Flatness (Unlock ed)

11. Mov e the mixer's IF output from the DVM to the sp ectrum analyzer's 50 input (D).
12. On the syn thesizer, turn on FM DC. Press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FM On/Off DC (asterisk on)

MOD

13. On the sp ectrum analyzer, set:
Start Frequency100 kHz
Stop Frequency10 MHz
dB/Div1 dB

Change the reference lev el to set the start of the trace or the highest p eak on the cen ter

graticule.

14. The display no w sho ws the syn thesizer FM atness from 100 kHz to 10 MHz. Use

the sp ectrum analyzer markers to nd the 3 dB bandwidth. Record this frequency in

T able 2-28. If the total v ariation is less than 3 dB then record \ > 10 MHz" in T able 2-28.

15. Repeat steps 7 through 14 at the syn thesizer frequencies in T able 2-28.
16. Record the smallest bandwidth from T able 2-28 on the test record.

In Case of Difficulty

Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

Agilent 8360 P erformance T ests

Show 5 more labels / short text on this page
· · Related Adjustments · FM Gain · FM Bandwidth
Page 189 of the Agilent Technologies 83640B user manual — CW
Page 189 · click to enlarge
Page 189
21. Maximum FM Deviation

(F or B-Series Only)

In this pro cedure, the syn thesizer's RF output is monitored directly on a sp ectrum analyzer

for low FM rates. The FM amplitude is increased un til the maximum deviation exceeds the

sp eci cation.

F or higher FM rates, the FM rate is set so that a Bessel n ull o ccurs when the RF p eak

deviation equals the sp eci cation. The mo dulation amplitude is increased in order to pass

through the correct Bessel n ull for the FM rate used and the p eak deviation sp eci ed.

1. Connect the equipmen t as sho wn in Figure 2-30. Preset all the instrumen ts and let them

w arm up for at least one hour.

Agilent Technologies 83640B — 08360125.pdf diagram (page 189)
Fig. 37 · 08360125.pdf

Figure 2-30. Maximum FM Deviation T est Setup

Sp ectrum Analyzer HP/Agilent 8566B

F unction Generator HP/Agilent 3325A

F unction Generator HP/Agilent 8111A or HP/Agilent 8116A

DC FM (100 Hz) (Using HP/Agilent 8111A or HP/Agilent 8116A)

Syn thesizer mo dels without 1 GHz capabilit y should use 5 GHz settings

instead of 1 GHz, as indicated in paren thesis.

2. On the syn thesizer, press:

CW

2-100 P erformance T ests

Show 8 more labels / short text on this page
Description and Procedure · Equipment · · · 1 · GHz · (5 GHz) · Maximum FM Deviation
Page 190 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 190 · click to enlarge
Page 190

FREQUENCY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Up/Dn Size CW

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USER CAL

Peak RF Always (asterisk on)

3. On the sp ectrum analyzer, set:
Center Frequency1 GHz (5 GHz)
Frequency Span187.5 MHz
Reference Level10 dBm
4. Connect the high v oltage output (16 V p-p) function generator and set it as follows:
FunctionSine wave
Frequency100 Hz
AmplitudeMinimum
5. Cen ter the signal on the sp ectrum analyzer display b y adjusting the cen ter frequency . Set

the sp ectrum analyzer to maximum hold.

6. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MOD

FM On/Off DC (asterisk on)

or for syn thesizers with Option 002, press:

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FM On/Off Ext (asterisk on)

MOD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FM Coupling DC (asterisk on)

The function generator m ust b e capable of supplying > 16 Vp-p in to 50 to

obtain sucien t FM deviation for this step.

7. Increase the function generator amplitude un til the width of the signal display ed exceeds

eigh t divisions on the sp ectrum analyzer. In T able 2-29 record P ASS if the signal exceeds

eigh t divisions with no unlo c k or ov ermo d message on the syn thesizer. Record F AIL if

not.

T able 2-29. DC FM T est P ass-F ail Results

1 Disregard if b eyond your synthesizer's capability .

8. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FM On/Off DC (asterisk o )

or for syn thesizers with Option 002, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FM On/Off Ext (asterisk o )

9. On the sp ectrum analyzer, clear and write to trace A.
10. Decrease the function generator amplitude to minimum.
11. Repeat steps 5 through 10 for the remaining syn thesizer and sp ectrum analyzer

frequencies in T able 2-29 .

Agilent 8360 P erformance T ests

Show 18 more labels / short text on this page
· 1 · · GHz · · · Tracking Menu · · · · · FM Menu · 1 GHz · 5 GHz · 9 GHz · 18 GHz · 22 GHz · Maximum FM Deviation
Page 191 of the Agilent Technologies 83640B user manual — AMPTD CAL
Page 191 · click to enlarge
Page 191
12. Circle P ASS on the test record if all the results in T able 2-29 are P ASS.

If not, circle F AIL on the test record.

A C FM (at 100 kHz) (Using HP/Agilent 3325A)

Maxim um FM deviation for a 100 kHz rate is limited to a mo dulation index of

5 times n, where n = frequency band. The mo dulation index is equiv alen t to

deviation/rate.

Syn thesizer mo dels without 1 GHz capabilit y should use 5 GHz settings, as

indicated in paren thesis, instead of 1 GHz.

13. On the syn thesizer, press:

CW

14. On the sp ectrum analyzer, set:
Center frequencySame as synthesizer frequency
Frequency Span1 MHz
Reference Level10 dBm

Clear all data from the display.

15. Cen ter the signal on the analyzer display .
16. Connect the syn thesized function generator and set it as follows:
FunctionSine wave
Frequency (FM Rate)100 kHz
Amplitude1 mV
17. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FM On/Off 100 kHz (asterisk on)

MOD

or, for syn thesizers with Option 002, press:

MOD

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FM On/Off Ext (asterisk on)

This activ ates FM sensitivity . Use the do wn arrow k ey to set FM sensitivity to 1.00

MHz/V.

18. On the function generator, set:
Amplitude (Calculated Null Voltage)271 mV rms

k ey to do an amplitude calibration. Use the arrow k eys to adjust the

AMPTD CAL

amplitude to ac hiev e a n ull of the 1st sideband (Sideband Nulled).

The sidebands may not resp ond to increases and decreases of the FM input

v oltage symmetrically . In this case, take the av erage of the v oltages that cause

the low er sideband and upp er sideband to n ull. If the sidebands n ull at more

than one v oltage lev el, use the low er v oltage v alue as the n ull v oltage.

19. Record the \Actual Null V oltage" in the space provided in T able 2-30.
20. Calculate the actual FM sensitivit y using the formula:

Calculated Null V oltage

Actual FM Sensitivity =

Actual Null V oltage

Record this v alue in the space provided in T able 2-30.

2-102 P erformance T ests

Show 15 more labels / short text on this page
· · 1 · GHz · (5 GHz) · · · · · FM Menu · Use the · · · (Set FM Sensitivity) · Maximum FM Deviation
Page 192 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 192 · click to enlarge
Page 192
21. On the function generator, increase the amplitude (FM input v oltage) un til the

syn thesizer displays an UNLOCK or O VERMOD message. Decrease the FM Input

V oltage just un til the UNLOCK or O VERMOD message turns o .

22. Record the FM input v oltage in the Maximum FM V oltage column of T able 2-30.
23. Calculate the maximum FM deviation using the formula:

Max FM Deviation = (Actual FM Sensitivit y)[(1.414)(Max FM V oltage)]

24. Record the maximum FM deviation in the appropriate column of the test record.
25. Repeat steps 13 through 24 for all the syn thesizer settings indicated in T able 2-30.

Remem b er to use the set FM sensitivit y as a m ultiplier when calculating the

actual FM sensitivit y .

T able 2-30. FM Deviation Frequencies and Settings

Agilent Technologies 83640B — 08360125.pdf diagram (page 192)
Fig. 38 · 08360125.pdf

1 Disregard if b eyond your synthesizer's capabilit y .

Agilent 8360 P erformance T ests

Maximum FM Deviation

Page 193 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 193 · click to enlarge
Page 193

Mo dulation Generator Flatness

In Case of Difficulty

1. If the syn thesizer remains unlo c k ed, turn A C FM o and b egin the pro cedure from step 13

for the CW frequency in question.

2. Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

2-104 P erformance T ests

Show 4 more labels / short text on this page
Related Adjustments · FM Gain · Mo dulation Generator · Maximum FM Deviation
Page 194 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 194 · click to enlarge
Page 194
22. Internal Pulse Accuracy (Option 002)

(F or B-Series Only)

This p erformance test uses an oscilloscop e to measure the pulse width of the output of the

mo dulation generator (not the pulse width of the RF output).

1. Connect the equipmen t as sho wn in Figure 2-31. Preset all instrumen ts and let them w arm

up for at least one hour.

Agilent Technologies 83640B — 08360125.pdf diagram (page 194)
Fig. 39 · 08360125.pdf

Figure 2-31. Pulse Accuracy T est Setup

Digitizing Oscilloscop e HP/Agilent 54111D

Oscilloscop e Prob es HP/Agilent 10437A

2. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Pulse On/Off Int (asterisk on)

MOD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Internal Pulse Width

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Internal Pulse Period

Agilent 8360 P erformance T ests

Show 13 more labels / short text on this page
Description and Procedure · Equipment · · · Pulse Menu · Internal Menu · 25 · ns · · 300 · ns · · Internal Pulse Accuracy
Page 195 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 195 · click to enlarge
Page 195
3. On the oscilloscop e, set:

Channel 1:

Timebase:

Trigger:

Trig Src Chan 1

Trigger Level 1.2V

Display:

Display Time 0.2 s

4. Adjust the oscilloscop e o set to fully display the signal.
5. Use the oscilloscop e to measure the width of the pulse. Record the v alue measured on the

test record.

6. On the syn thesizer, c hange the pulse width to 50 ns. Press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Internal Pulse Width

7. Use the oscilloscop e to measure the width of the pulse. Record the v alue measured on the

test record.

In Case of Difficulty

Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

2-106 P erformance T ests

Show 17 more labels / short text on this page
Display On · Volts/Division 500 mV · Input Coupling dc · Input Impedance 50 · Time/Division 10 ns · Delay Reference center · Sweep auto · Trigger Mode Edge · Slope Pos · Display Mode Repetitive · Averaging Off · 50 · · ns · Related Adjustments · None · Internal Pulse Accuracy
Page 196 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 196 · click to enlarge
Page 196
23. Modulation Meter (Option 002)

(F or B-Series Only)

The mo dulation meter measures and displays the v alue of the externally generated amplitude

or frequency mo dulation. This pro cedure determines the accuracy of the mo dulation meter.

1. Connect the equipmen t as sho wn in Figure 2-32. Preset the instrumen ts and let them

w arm up for at least one hour.

Agilent Technologies 83640B — 08360125.pdf diagram (page 196)
Fig. 40 · 08360125.pdf

Figure 2-32. Modulation Meter T est Setup

F unction Generator HP/Agilent 3325A

Digital V oltmeter HP/Agilent 3456A

2. Set the DVM to measure A C. Press:

STORE

N CYC INT

STORE

N DIG DISP

3. On the syn thesizer, press

CW

.

4. Set the function generator for a 100 kHz, 600 mV p-p, sine w av e output signal.
5. On the syn thesizer, turn on external AM mo dulation and turn on the

AM meter function. Press:

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM On/Off Ext (asterisk on).

MOD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Meter On/Off AM (asterisk on).

PRIOR

6. Record the AM METER v alue on the test record.
7. Set the function generator for a 1 kHz, 1.800 V p-p, sine w av e output signal. If necessary ,

adjust the function generator for a DVM reading of 0.6363 A C V rms.

8. Record the AM METER v alue on the test record.
9. Disconnect the BNC tee from the AM input and connect it to the FM input.
10. On the syn thesizer, turn o external AM, turn on external FM, set the FM sensitivity to

1 MHz/V, and turn on the FM meter function. Press:

MOD

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM On/Off Ext (asterisk o )

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRIOR

FM On/Off Ext (asterisk on)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRIOR

Meter On/Off FM (asterisk on)

Agilent 8360 P erformance T ests

Show 33 more labels / short text on this page
Description and Procedure · Equipment · · 10 · · · · 6 · · · · 10 · GHz · · · AM Menu · · · Monitor Menu · · · AM Menu · · · FM Menu · · 1 · · MHz · · · Monitor Menu · Modulation Meter
Page 197 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 197 · click to enlarge
Page 197
11. Set the function generator for a 1 kHz, 10 V p-p, sine w av e output signal. If necessary ,

adjust the function generator amplitude for a DVM reading of 3.535 A C V rms.

12. Record the FM METER v alue on the test record.
13. Set the function generator for a 100 kHz, 2 V p-p, sine w av e output signal.
14. Record the FM METER v alue on the test record.

In Case of Difficulty

Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

2-108 P erformance T ests

Related Adjustments · None · Modulation Meter

Page 198 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 198 · click to enlarge
Page 198

T able 2-33. T est Record for Agilent 83620B and 83622B (1 of 8)

T est F acilityRep ort Num b er

Mo del Am bien t temp erature

Serial Num b er Relative humidit y %

Options Line frequency Hz (nominal)

Firm w are Revision

Sp ecial Notes:

Agilent 8360 P erformance T ests

Agilent 83620B/22B T est Record

Show 4 more labels / short text on this page
Date · Customer · T ested by ·
Page 199 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 199 · click to enlarge
Page 199

T able 2-33. T est Record for 83620B and 83622B (2 of 8)

Mo del Rep ort Num b er Date

T est Equipmen t Used Mo del Numb er T race Numb er Cal Due Date

1. Digital Oscilloscop e
2. Measuring Receiver
3. P ow er Sensor
4. P ow er Meter
5. Microw ave Sp ectrum Analyzer
6. F unction Generator

2-110 P erformance T ests

Agilent 83620B/22B T est Record

Page 200 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 200 · click to enlarge
Page 200

T able 2-33. T est Record for 83620B and 83622B (3 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 200)
Fig. 41 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertain t y

10. Start F req 0.1% of 0.005% of
8. CW Step 50 ms 10 s
16. Stepp ed Sw eep 5.5 ms 1 s
25. F requency List 50 ms 10 s

< 2 GHz 0.6 dB +0.6 dB 0.19 dB

2 and 20 GHz 0.7 dB +0.7 dB 0.35 dB

< 2 GHz 1.2 dB +1.2 dB 0.21 dB

2 and 20 GHz 1.3 dB +1.3 dB 0.46 dB

2 and 20 GHz 1.2 dB 0.25 dB

6. 20 GHz +13 dBm 1.8 dB
6. 20 GHz +11.5 dBm 1.8 dB

Agilent 8360 P erformance T ests

Agilent 83620B/22B T est Record

Page 201 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 201 · click to enlarge
Page 201

T able 2-33. T est Record for 83620B and 83622B (4 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 201)
Fig. 42 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertaint y

7. Minim um P ow er 0.2 mV 3 V
10. Maximum P ow er 0.5 V 20 V
8. 0.01 to < 2.0 GHz 30 dBc 1.75 dB
12. 2.0 and 20 GHz 50 dBc 1.75 dB
8. < 50 MHz 20 dBc 1.75 dB
8. 50 MHz to < 2.0 GHz 30 dBc 1.75 dB
12. 2.0 and 20 GHz 60 dBc 1.75 dB
12. 7 and 20 GHz 50 dBc 1.75 dB

100 MHz Fixed 60 dBc 1 dB

100 MHz O set 60 dBc 1 dB

LO F eedthrough 60 dBc 1 dB

20. Low Band Mixer Spur 60 dBc 1.85 dB

2-112 P erformance T ests

Agilent 83620B/22B T est Record

Page 202 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 202 · click to enlarge
Page 202

T able 2-33. T est Record for 83620B and 83622B (5 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 202)
Fig. 43 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertain t y

120 Hz Spur 55 dBc 1 dB

180 Hz Spur 55 dBc 1 dB

240 Hz Spur 55 dBc 1 dB

120 Hz Spur 55 dBc 1 dB

180 Hz Spur 55 dBc 1 dB

240 Hz Spur 55 dBc 1 dB

100 Hz to < 1 kHz 70 dBc 2 dB

1 kHz to < 10 kHz 78 dBc 2 dB

10 kHz to < 100 kHz 86 dBc 2 dB

100 kHz 107 dBc 2 dB

100 Hz to < 1 kHz 70 dBc 2 dB

1 kHz to < 10 kHz 78 dBc 2 dB

10 kHz to < 100 kHz 86 dBc 2 dB

100 kHz 107 dBc 2 dB

100 Hz to < 1 kHz 60 dBc 2 dB

1 kHz to < 10 kHz 68 dBc 2 dB

10 kHz to < 100 kHz 76 dBc 2 dB

100 kHz 97 dBc 2 dB

1 3 dB if the Invertron is not used.

Agilent 8360 P erformance T ests

Agilent 83620B/22B T est Record

Page 203 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 203 · click to enlarge
Page 203

T able 2-33. T est Record for 83620B and 83622B (6 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 203)
Fig. 44 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertaint y

21. Level Accuracy 0.3 dB 0.1 dB
21. Level Accuracy 0.3 dB 0.1 dB

2-114 P erformance T ests

Agilent 83620B/22B T est Record

Page 204 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 204 · click to enlarge
Page 204

T able 2-33. T est Record for 83620B and 83622B (7 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 204)
Fig. 45 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertain t y

11. AM Depth Opt. 002 28.5% 31.5% 0.76%
21. AM Depth Opt. 002 28.5% 31.5% 0.76%
10. 3 dB Bandwidth 100 kHz 10.31 kHz
13. Maximum FM Error 10% +10% 3.7%
23. Maximum FM Error Opt. 002 10% +10% 3.7%
16. 3 dB Bandwidth 8 MHz 320 kHz

Agilent 8360 P erformance T ests

Agilent 83620B/22B T est Record

Page 205 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 205 · click to enlarge
Page 205

T able 2-33. T est Record for 83620B and 83622B (8 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 205)
Fig. 46 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertaint y

> 75 MHz Deviation P ass/F ail 2.3 MHz

5. Pulse Width 20 ns 30 ns 110 ps
7. Pulse Width 45 ns 55 ns 117 ps
11. FM Meter 4.95 MHz 5.05 MHz 15 kHz
13. FM Meter 950 kHz 1.05 MHz 10 kHz

2-116 P erformance T ests

Agilent 83620B/22B T est Record

Page 206 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 206 · click to enlarge
Page 206

T able 2-34. T est Record for Agilent 83623B and 83624B (1 of 8)

T est F acilityRep ort Num b er

Mo del Am bien t temp erature

Serial Num b er Relative humidit y %

Options Line frequency Hz (nominal)

Firm w are Revision

Sp ecial Notes:

Agilent 8360 P erformance T ests

Agilent 83623B/24B T est Record

Show 4 more labels / short text on this page
Date · Customer · T ested by ·
Page 207 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 207 · click to enlarge
Page 207

T able 2-34. T est Record for 83623B and 83624B (2 of 8)

Mo del Rep ort Num b er Date

T est Equipmen t Used Mo del Numb er T race Numb er Cal Due Date

1. Digital Oscilloscop e
2. Measuring Receiver
3. P ow er Sensor
4. P ow er Meter
5. Microw ave Sp ectrum Analyzer
6. F unction Generator

2-118 P erformance T ests

Agilent 83623B/24B T est Record

Page 208 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 208 · click to enlarge
Page 208

T able 2-34. T est Record for 83623B and 83624B (3 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 208)
Fig. 47 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertain t y

10. Start F req 0.1% of 0.005% of
8. CW Step 50 ms 10 s
16. Stepp ed Sw eep 5.5 ms 1 s
25. F requency List 50 ms 10 s

< 2 GHz 0.6 dB +0.6 dB 0.19 dB

2 and 20 GHz 0.7 dB +0.7 0.35 dB

< 2 GHz 1.2 dB +1.2 dB 0.21 dB

2 and 20 GHz 1.3 dB +1.3 dB 0.46 dB

2 and 20 GHz 1.2 dB 0.25 dB

Standard or Opt. 006 +17 dBm 1.8 dB

Opt. 001 or Opts. 001 and 006 +15.5 dBm 1.8 dB

Agilent 8360 P erformance T ests

Agilent 83623B/24B T est Record

Page 209 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 209 · click to enlarge
Page 209

T able 2-34. T est Record for 83623B and 83624B (4 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 209)
Fig. 48 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertaint y

Opts. 001 and 006 +15.5 dBm 1.8 dB

7. Minim um P ow er 0.2 mV 3 V
10. Maximum P ow er 0.5 V 20 V
8. < 50 MHz 20 dBc 1.75 dB
8. 50 MHz to < 2.0 GHz 25 dBc 1.75 dB
12. 2.0 to 20 GHz 25 dBc 1.75 dB
8. < 50 MHz 20 dBc 1.75 dB
8. 50 MHz to < 2.0 GHz 25 dBc 1.75 dB
12. 2.0 and 20 GHz 60 dBc 1.75 dB
12. 7 to 20 GHz 50 dBc 1.75 dB

2-120 P erformance T ests

Agilent 83623B/24B T est Record

Page 210 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 210 · click to enlarge
Page 210

T able 2-34. T est Record for 83623B and 83624B (5 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 210)
Fig. 49 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertain t y

100 MHz Fixed 60 dBc 1 dB

100 MHz O set 60 dBc 1 dB

LO F eedthrough 60 dBc 1 dB

20. Low Band Mixer Spur 60 dBc 1.85 dB

120 Hz Spur 55 dBc 1 dB

180 Hz Spur 55 dBc 1 dB

240 Hz Spur 55 dBc 1 dB

120 Hz Spur 55 dBc 1 dB

180 Hz Spur 55 dBc 1 dB

240 Hz Spur 55 dBc 1 dB

100 Hz to < 1 kHz 70 dBc 2 dB

1 kHz to < 10 kHz 78 dBc 2 dB

10 kHz to < 100 kHz 86 dBc 2 dB

100 kHz 107 dBc 2 dB

100 Hz to < 1 kHz 70 dBc 2 dB

1 kHz to < 10 kHz 78 dBc 2 dB

10 kHz to < 100 kHz 86 dBc 2 dB

100 kHz 107 dBc 2 dB

1 3 dB if the Invertron is not used.

Agilent 8360 P erformance T ests

Agilent 83623B/24B T est Record

Page 211 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 211 · click to enlarge
Page 211

T able 2-34. T est Record for 83623B and 83624B (6 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 211)
Fig. 50 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertaint y

100 Hz to < 1 kHz 60 dBc 2 dB

1 kHz to < 10 kHz 68 dBc 2 dB

10 kHz to < 100 kHz 76 dBc 2 dB

100 kHz 97 dBc 2 dB

21. Level Accuracy 0.3 dB 0.1 dB
21. Level Accuracy 0.3 dB 0.1 dB
26. 2 and 20.0 GHz 1% 0.06%

2-122 P erformance T ests

Agilent 83623B/24B T est Record

Page 212 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 212 · click to enlarge
Page 212

T able 2-34. T est Record for 83623B and 83624B (7 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 212)
Fig. 51 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertain t y

11. AM Depth Opt. 002 28.5% 31.5% 0.76%
21. AM Depth Opt. 002 28.5% 31.5% 0.76%
10. 3 dB Bandwidth 100 kHz 10.31 kHz
13. Maximum FM Error 10% +10% 3.7%
23. Maximum FM Error Opt. 002 10% +10% 3.7%
16. 3 dB Bandwidth 8 MHz 320 kHz

Agilent 8360 P erformance T ests

Agilent 83623B/24B T est Record

Page 213 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 213 · click to enlarge
Page 213

T able 2-34. T est Record for 83623B and 83624B (8 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 213)
Fig. 52 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertaint y

> 75 MHz Deviation P ass/F ail 2.3 MHz

5. Pulse Width 20 ns 30 ns 110 ps
7. Pulse Width 45 ns 55 ns 117 ps
11. FM Meter 4.95 MHz 5.05 MHz 15 kHz
13. FM Meter 950 kHz 1.05 MHz 10 kHz

2-124 P erformance T ests

Agilent 83623B/24B T est Record

Page 214 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 214 · click to enlarge
Page 214

T able 2-35. T est Record for Agilent 83630B (1 of 8)

T est F acilityRep ort Num b er

Mo del Am bien t temp erature

Serial Num b er Relative humidit y %

Options Line frequency Hz (nominal)

Firm w are Revision

Sp ecial Notes:

Agilent 8360 P erformance T ests

Agilent 83630B T est Record

Show 4 more labels / short text on this page
Date · Customer · T ested by ·
Page 215 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 215 · click to enlarge
Page 215

T able 2-35. T est Record for 83630B (2 of 8)

Mo del Rep ort Num b er Date

T est Equipmen t Used Mo del Numb er T race Numb er Cal Due Date

1. Digital Oscilloscop e
2. Measuring Receiver
3. P ow er Sensor
4. P ow er Meter
5. Microw ave Sp ectrum Analyzer
6. F unction Generator

2-126 P erformance T ests

Agilent 83630B T est Record

Page 216 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 216 · click to enlarge
Page 216

T able 2-35. T est Record for 83630B (3 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 216)
Fig. 53 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertain t y

10. Start F req 0.1% of 0.005% of
8. CW Step 50 ms 10 s
16. Stepp ed Sw eep 5.5 ms 1 s
25. F requency List 50 ms 10 s

< 2 GHz 0.6 dB +0.6 dB 0.19 dB

2 and 20 GHz 0.7 dB +0.7 dB 0.35 dB

> 20 and 26.5 GHz 0.9 dB +0.9 dB 0.38 dB

< 2 GHz 1.2 dB +1.2 dB 0.21 dB

2 and 20 GHz 1.3 dB +1.3 dB 0.46 dB

2 and 20 GHz 1.2 dB 0.25 dB

> 20 and 26.5 GHz 1.6 dB 0.27 dB

0.01 to 26.5 GHz +10.0 dBm 1.7 dB

0.01 to 20 GHz +13.0 dBm 1.7 dB

Agilent 8360 P erformance T ests

Agilent 83630B T est Record

Page 217 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 217 · click to enlarge
Page 217

T able 2-35. T est Record for 83630B (4 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 217)
Fig. 54 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertaint y

0.01 to 26.5 GHz +8.0 dBm 1.7 dB

0.01 to 20 GHz +11.5 dBm 1.7 dB

7. Minim um P ow er 0.2 mV 3 V
10. Maximum P ow er 0.5 V 20 V
8. 0.01 to < 2.0 GHz 30 dBc 1.75 dB
21. 2.0 and 26.5 GHz 50 dBc 1.75 dB
8. < 50 MHz 20 dBc 1.75 dB
8. 50 MHz to < 2.0 GHz 30 dBc 1.75 dB
12. 2.0 and 26.5 GHz 60 dBc 1.75 dB
12. 7 and 20 GHz 50 dBc 1.75 dB
21. > 20 and 26.5 GHz 50 dBc 1.75 dB

2-128 P erformance T ests

Agilent 83630B T est Record

Page 218 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 218 · click to enlarge
Page 218

T able 2-35. T est Record for 83630B (5 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 218)
Fig. 55 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertain t y

100 MHz Fixed 60 dBc 1 dB

100 MHz O set 60 dBc 1 dB

LO F eedthrough 60 dBc 1 dB

20. Low Band Mixer Spur 60 dBc 1.85 dB

120 Hz Spur 55 dBc 1 dB

180 Hz Spur 55 dBc 1 dB

240 Hz Spur 55 dBc 1 dB

120 Hz Spur 55 dBc 1 dB

180 Hz Spur 55 dBc 1 dB

240 Hz Spur 55 dBc 1 dB

100 Hz to < 1 kHz 70 dBc 2 dB

1 kHz to < 10 kHz 78 dBc 2 dB

10 kHz to < 100 kHz 86 dBc 2 dB

100 kHz 107 dBc 2 dB

100 Hz to < 1 kHz 70 dBc 2 dB

1 kHz to < 10 kHz 78 dBc 2 dB

10 kHz to < 100 kHz 86 dBc 2 dB

100 kHz 107 dBc 2 dB

1 3 dB if the Invertron is not used.

Agilent 8360 P erformance T ests

Agilent 83630B T est Record

Page 219 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 219 · click to enlarge
Page 219

T able 2-35. T est Record for 83630B (6 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 219)
Fig. 56 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertaint y

100 Hz to < 1 kHz 60 dBc 2 dB

1 kHz to < 10 kHz 68 dBc 2 dB

10 kHz to < 100 kHz 76 dBc 2 dB

100 kHz 97 dBc 2 dB

21. Level Accuracy 0.3 dB 0.1 dB
21. Level Accuracy 0.3 dB 0.3 dB

2-130 P erformance T ests

Agilent 83630B T est Record

Page 220 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 220 · click to enlarge
Page 220

T able 2-35. T est Record for 83630B (7 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 220)
Fig. 57 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertain t y

11. AM Depth Opt. 002 28.5% 31.5% 0.76%
21. AM Depth Opt. 002 28.5% 31.5% 0.76%
10. 3 dB Bandwidth 100 kHz 10.31 kHz
23. Deep < 20 GHz 37 dBm 1.65 dB
23. Deep 20 GHz 40 dBm 1.35 dB
13. Maximum FM Error 10% +10% 3.7%
23. Maximum FM Error Opt. 002 10% +10% 3.7%
16. 3 dB Bandwidth 8 MHz 320 kHz

Agilent 8360 P erformance T ests

Agilent 83630B T est Record

Page 221 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 221 · click to enlarge
Page 221

T able 2-35. T est Record for 83630B (8 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 221)
Fig. 58 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertaint y

> 75 MHz Deviation P ass/F ail 2.3 MHz

5. Pulse Width 20 ns 30 ns 110 ps
7. Pulse Width 45 ns 55 ns 117 ps
11. FM Meter 4.95 MHz 5.05 MHz 15 kHz
13. FM Meter 950 kHz 1.05 MHz 10 kHz

2-132 P erformance T ests

Agilent 83630B T est Record

Page 222 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 222 · click to enlarge
Page 222

T able 2-36. T est Record for Agilent 83640B (1 of 8)

T est F acilityRep ort Num b er

Mo del Am bien t temp erature

Serial Num b er Relative humidit y %

Options Line frequency Hz (nominal)

Firm w are Revision

Sp ecial Notes:

Agilent 8360 P erformance T ests

Agilent 83640B T est Record

Show 4 more labels / short text on this page
Date · Customer · T ested by ·
Page 223 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 223 · click to enlarge
Page 223

T able 2-36. T est Record for 83640B (2 of 8)

Mo del Rep ort Num b er Date

T est Equipmen t Used Mo del Numb er T race Numb er Cal Due Date

1. Digital Oscilloscop e
2. Measuring Receiver
3. P ow er Sensor
4. P ow er Meter
5. Microw ave Sp ectrum Analyzer
6. F unction Generator

2-134 P erformance T ests

Agilent 83640B T est Record

Page 224 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 224 · click to enlarge
Page 224

T able 2-36. T est Record for 83640B (3 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 224)
Fig. 59 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertain t y

10. Start F req 0.1% of 0.005% of
8. CW Step 50 ms 10 s
16. Stepp ed Sw eep 5.5 ms 1 s
25. F requency List 50 ms 10 s

< 2 GHz 0.6 dB 0.6 dB 0.22 dB

2 and 20 GHz 0.7 dB 0.7 dB 0.35 dB

> 20 and 40 GHz 0.9 dB 0.9 dB 0.38 dB

2 and 20 GHz 1.2 dB 0.25 dB

> 20 and 40 GHz 1.8 dB 0.27 dB

0.01 to 40 GHz +6.0 dBm 1.4 dB

0.01 to < 26.5 GHz +10.0 dBm 1.8 dB

0.01 to < 20 GHz +10.0 dBm 1.8 dB

Agilent 8360 P erformance T ests

Agilent 83640B T est Record

Page 225 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 225 · click to enlarge
Page 225

T able 2-36. T est Record for the 83640B (4 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 225)
Fig. 60 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertaint y

0.01 to 40 GHz +4 dBm 1.4 dB

0.01 to < 26.5 GHz +8 dBm 1.8 dB

0.01 to < 20 GHz +8.5 dBm 1.8 dB

7. Minim um P ow er 0.2 mV 3 V
10. Maximum P ow er 0.5 V 20 V
8. < 50 MHz 20 dBc 1.75 dB
8. 50 MHz and < 2.0 GHz 30 dBc 1.75 dB
21. 2.0 and < 26.5 GHz 50 dBc 1.75 dB
29. 26.5 to 40 GHz 40 dBc 1.75 dB
12. 7 and 20 GHz 50 dBc 1.75 dB
29. > 20 and 40 GHz 40 dBc 1.55 dB

100 MHz Fixed 60 dBc 1 dB

100 MHz O set 60 dBc 1 dB

LO F eedthrough 60 dBc 1 dB

2-136 P erformance T ests

Agilent 83640B T est Record

Page 226 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 226 · click to enlarge
Page 226

T able 2-36. T est Record for the 83640B (5 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 226)
Fig. 61 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertain t y

20. Low Band Mixer Spur 60 dBc 1.85 dB

100 Hz to < 1 kHz 70 dBc 2 dB

1 kHz to < 10 kHz 78 dBc 2 dB

10 kHz to < 100 kHz 86 dBc 2 dB

100 kHz 107 dBc 2 dB

100 Hz to < 1 kHz 70 dBc 2 dB

1 kHz to < 10 kHz 78 dBc 2 dB

10 kHz to < 100 kHz 86 dBc 2 dB

100 kHz 107 dBc 2 dB

100 Hz to < 1 kHz 60 dBc 2 dB

1 kHz to < 10 kHz 68 dBc 2 dB

10 kHz to < 100 kHz 76 dBc 2 dB

100 kHz 97 dBc 2 dB

1 3 dB if the Invertron is not used.

Agilent 8360 P erformance T ests

Agilent 83640B T est Record

Page 227 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 227 · click to enlarge
Page 227

T able 2-36. T est Record for the 83640B (6 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 227)
Fig. 62 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertaint y

21. Level Accuracy 0.3 dB 0.1 dB
21. Level Accuracy 0.3 dB 0.1 dB
26. 2 and 40.0 GHz 1% 0.06%

2-138 P erformance T ests

Agilent 83640B T est Record

Page 228 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 228 · click to enlarge
Page 228

T able 2-36. T est Record for the 83640B (7 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 228)
Fig. 63 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertain t y

11. AM Depth Opt. 002 28.5% 31.5% 0.76%
21. AM Depth Opt. 002 28.5% 31.5% 0.76%
32. AM Depth Opt. 002 28.5% 31.5% 0.76%
10. 3 dB Bandwidth 100 kHz 10.31 kHz
23. Deep 20 GHz 40 dBm 1.65 dB
23. Deep > 20 GHz 30 dBm 1.35 dB
13. Maximum FM Error 10% 10% 3.7%
23. Maximum FM Error Opt. 002 10% 10% 3.7%
16. 3 dB Bandwidth 8 MHz 320 kHz

Agilent 8360 P erformance T ests

Agilent 83640B T est Record

Page 229 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 229 · click to enlarge
Page 229

T able 2-36. T est Record for the 83640B (8 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 229)
Fig. 64 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertaint y

> 75 MHz Deviation P ass/F ail 2.3 MHz

5. Pulse Width 20 ns 30 ns 110 ps
7. Pulse Width 45 ns 55 ns 117 ps
11. FM Meter 4.95 MHz 5.05 MHz 15 kHz
13. FM Meter 950 kHz 1.05 MHz 10 kHz

2-140 P erformance T ests

Agilent 83640B T est Record

Page 230 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 230 · click to enlarge
Page 230

T able 2-37. T est Record for Agilent 83650B (1 of 8)

T est F acilityRep ort Num b er

Mo del Am bien t temp erature

Serial Num b er Relative humidit y %

Options Line frequency Hz (nominal)

Firm w are Revision

Sp ecial Notes:

Agilent 8360 P erformance T ests

Agilent 83650B T est Record

Show 4 more labels / short text on this page
Date · Customer · T ested by ·
Page 231 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 231 · click to enlarge
Page 231

T able 2-37. T est Record for 83650B (2 of 8)

Mo del Rep ort Num b er Date

T est Equipmen t Used Mo del Numb er T race Numb er Cal Due Date

1. Digital Oscilloscop e
2. Measuring Receiver
3. P ow er Sensor
4. P ow er Meter
5. Microw ave Sp ectrum Analyzer
6. F unction Generator

2-142 P erformance T ests

Agilent 83650B T est Record

Page 232 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 232 · click to enlarge
Page 232

T able 2-37. T est Record for 83650B (3 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 232)
Fig. 65 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertain t y

10. Start F req 0.1% of 0.005% of
8. CW Step: 50 ms 10 s
16. Stepp ed Sw eep 5.5 ms 1 s
25. F requency List 50 ms 10 s

< 2 GHz 0.6 dB 0.6 dB 0.22 dB

2 and 20 GHz 0.7 dB 0.7 dB 0.35 dB

> 20 and 40 GHz 0.9 dB 0.9 dB 0.38 dB

> 40 GHz 1.7 dB 1.7 dB 0.79 dB

2 and 20 GHz 1.2 dB 0.25 dB

> 20 and 40 GHz 1.6 dB 0.38 dB

0.01 to 50 GHz +2.5 dBm 2.3 dB

0.01 to < 40 GHz +5.0 dBm 1.3 dB

0.01 to < 26.5 GHz +10.0 dBm 1.7 dB

0.01 to < 20 GHz +10.0 dBm 1.7 dB

Agilent 8360 P erformance T ests

Agilent 83650B T est Record

Page 233 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 233 · click to enlarge
Page 233

T able 2-37. T est Record for 83650B (4 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 233)
Fig. 66 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertaint y

0.01 to 50 GHz 0 dBm 2.3 dB

0.01 to < 40 GHz +3.0 dBm 1.3 dB

0.01 to < 26.5 GHz +8.0 dBm 1.7 dB

0.01 to < 20 GHz +8.5 dBm 1.7 dB

7. Minim um P ow er 0.2 mV 3 V
10. Maximum P ow er 0.5 V 20 V
8. < 50 MHz 20 dBc 1.75 dB
8. 50 MHz to < 2.0 GHz 30 dBc 1.75 dB
21. 2.0 to 26.5 GHz 50 dBc 1.75 dB
37. 26.5 to 50 GHz 40 dBc 2.22 dB
12. 7 to 20 GHz 50 dBc 1.75 dB
29. > 20 to 40 GHz 40 dBc 1.55 dB
37. > 40 to 50 GHz 35 dBc 2.22 dB

2-144 P erformance T ests

Agilent 83650B T est Record

Page 234 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 234 · click to enlarge
Page 234

T able 2-37. T est Record for 83650B (5 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 234)
Fig. 67 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertain t y

100 MHz Fixed 60 dBc 1 dB

100 MHz O set 60 dBc 1 dB

LO F eedthrough 60 dBc 1 dB

20. Low Band Mixer Spur 60 dBc 1.85 dB

120 Hz Spur 55 dBc 1 dB

180 Hz Spur 55 dBc 1 dB

240 Hz Spur 55 dBc 1 dB

120 Hz Spur 55 dBc 1 dB

180 Hz Spur 55 dBc 1 dB

240 Hz Spur 55 dBc 1 dB

100 Hz to < 1 kHz 70 dBc 2 dB

1 kHz to < 10 kHz 78 dBc 2 dB

10 kHz to < 100 kHz 86 dBc 2 dB

100 kHz 107 dBc 2 dB

100 Hz to < 1 kHz 70 dBc 2 dB

1 kHz to < 10 kHz 78 dBc 2 dB

10 kHz to < 100 kHz 86 dBc 2 dB

100 kHz 107 dBc 2 dB

1 3 dB if the Invertron is not used.

Agilent 8360 P erformance T ests

Agilent 83650B T est Record

Page 235 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 235 · click to enlarge
Page 235

T able 2-37. T est Record for 83650B (6 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 235)
Fig. 68 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertaint y

100 Hz to < 1 kHz 60 dBc 2 dB

1 kHz to < 10 kHz 68 dBc 2 dB

10 kHz to < 100 kHz 76 dBc 2 dB

100 kHz 97 dBc 2 dB

21. Level Accuracy 0.3 dB 0.1 dB
21. Level Accuracy 0.3 dB 0.1 dB
26. 2 and 50.0 GHz 1% 0.06%

2-146 P erformance T ests

Agilent 83650B T est Record

Page 236 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 236 · click to enlarge
Page 236

T able 2-37. T est Record for 83650B (7 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 236)
Fig. 69 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertain t y

11. AM Depth Opt. 002 28.5% 31.5% 0.76%
21. AM Depth Opt. 002 28.5% 31.5% 0.76%
32. AM Depth Opt. 002 28.5% 31.5% 0.76%
10. 3 dB Bandwidth 100 kHz 10.31 kHz
23. Deep 20 GHz 40 dBm 1.65 dB
23. Deep > 20 GHz 30 dBm 1.35 dB
13. Maximum FM Error 10% +10% 3.7%
23. Maximum FM Error Opt. 002 10% +10% 3.7%
16. 3 dB Bandwidth 8 MHz 320 kHz

Agilent 8360 P erformance T ests

Agilent 83650B T est Record

Page 237 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 237 · click to enlarge
Page 237

T able 2-37. T est Record for 83650B (8 of 8)

Agilent Technologies 83640B — 08360125.pdf diagram (page 237)
Fig. 70 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertaint y

> 75 MHz Deviation P ass/F ail 2.3 MHz

5. Pulse Width 20 ns 30 ns 110 ps
7. Pulse Width 45 ns 55 ns 117 ps
11. FM Meter 4.95 MHz 5.05 MHz 15 kHz
13. FM Meter 950 kHz 1.05 MHz 10 kHz

2-148 P erformance T ests

Agilent 83650B T est Record

Page 238 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 238 · click to enlarge
Page 238

T able 2-34. T est Record for Agilent 83623L (1 of 5)

T est F acilityRep ort Num b er

Mo del Am bien t temp erature

Serial Num b er Relative humidit y %

Options Line frequency Hz (nominal)

Firm w are Revision

Sp ecial Notes:

Agilent 8360 P erformance T ests

Agilent 83623L T est Record

Show 4 more labels / short text on this page
Date · Customer · T ested by ·
Page 239 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 239 · click to enlarge
Page 239

T able 2-34. T est Record for 83623L (2 of 5)

Mo del Rep ort Num b er Date

T est Equipmen t Used Mo del Numb er T race Numb er Cal Due Date

1. Digitizing Oscilloscop e
2. Measuring Receiver
3. P ow er Sensor
4. P ow er Meter
5. Microw ave Sp ectrum Analyzer

2-150 P erformance T ests

Agilent 83623L T est Record

Page 240 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 240 · click to enlarge
Page 240

T able 2-34. T est Record for 83623L (3 of 5)

Agilent Technologies 83640B — 08360125.pdf diagram (page 240)
Fig. 71 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertain t y

10. Start F req 0.1% of 0.005% of
8. CW Step 50 ms 10 s
16. Stepp ed Sw eep 5.5 ms 1 s
25. F requency List 50 ms 10 s

< 2 GHz 0.6 dB +0.6 dB 0.19 dB

2 and 20 GHz 0.7 dB +0.7 0.35 dB

< 2 GHz 1.2 dB +1.2 dB 0.21 dB

2 and 20 GHz 1.3 dB +1.3 dB 0.46 dB

2 and 20 GHz 1.2 dB 0.25 dB

Agilent 8360 P erformance T ests

Agilent 83623L T est Record

Page 241 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 241 · click to enlarge
Page 241

T able 2-34. T est Record for 83623L (4 of 5)

Agilent Technologies 83640B — 08360125.pdf diagram (page 241)
Fig. 72 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertaint y

7. Minim um P ow er 0.2 mV 3 V
10. Maximum P ow er 0.5 V 20 V
8. < 50 MHz 20 dBc 1.75 dB
8. 50 MHz to < 2.0 GHz 25 dBc 1.75 dB
12. 2.0 to 20 GHz 45 dBc 1.75 dB
12. 7 to 20 GHz 50 dBc 1.75 dB

100 MHz Fixed 60 dBc 1 dB

100 MHz O set 60 dBc 1 dB

LO F eedthrough 60 dBc 1 dB

20. Low Band Mixer Spur 60 dBc 1.85 dB

2-152 P erformance T ests

Agilent 83623L T est Record

Page 242 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 242 · click to enlarge
Page 242

T able 2-34. T est Record for 83623L (5 of 5)

Agilent Technologies 83640B — 08360125.pdf diagram (page 242)
Fig. 73 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertain t y

120 Hz Spur 55 dBc 1 dB

180 Hz Spur 55 dBc 1 dB

240 Hz Spur 55 dBc 1 dB

120 Hz Spur 55 dBc 1 dB

180 Hz Spur 55 dBc 1 dB

240 Hz Spur 55 dBc 1 dB

100 Hz to < 1 kHz 70 dBc 2 dB

1 kHz to < 10 kHz 78 dBc 2 dB

10 kHz to < 100 kHz 86 dBc 2 dB

100 kHz 107 dBc 2 dB

100 Hz to < 1 kHz 70 dBc 2 dB

1 kHz to < 10 kHz 78 dBc 2 dB

10 kHz to < 100 kHz 86 dBc 2 dB

100 kHz 107 dBc 2 dB

100 Hz to < 1 kHz 60 dBc 2 dB

1 kHz to < 10 kHz 68 dBc 2 dB

10 kHz to < 100 kHz 76 dBc 2 dB

100 kHz 97 dBc 2 dB

1 3 dB if the Invertron is not used.

Agilent 8360 P erformance T ests

Agilent 83623L T est Record

Page 243 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 243 · click to enlarge
Page 243

T able 2-35. T est Record for Agilent 83630L (1 of 5)

T est F acilityRep ort Num b er

Mo del Am bien t temp erature

Serial Num b er Relative humidit y %

Options Line frequency Hz (nominal)

Firm w are Revision

Sp ecial Notes:

2-154 P erformance T ests

Agilent 83630L T est Record

Show 4 more labels / short text on this page
Date · Customer · T ested by ·
Page 244 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 244 · click to enlarge
Page 244

T able 2-35. T est Record for 83630L (2 of 5)

Mo del Rep ort Num b er Date

T est Equipmen t Used Mo del Numb er T race Numb er Cal Due Date

1. Digitizing Oscilloscop e
2. Measuring Receiver
3. P ow er Sensor
4. P ow er Meter
5. Microw ave Sp ectrum Analyzer

Agilent 8360 P erformance T ests

Agilent 83630L T est Record

Page 245 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 245 · click to enlarge
Page 245

T able 2-35. T est Record for 83630L (3 of 5)

Agilent Technologies 83640B — 08360125.pdf diagram (page 245)
Fig. 74 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertaint y

10. Start F req 0.1% of 0.005% of
8. CW Step 50 ms 10 s
16. Stepp ed Sw eep 5.5 ms 1 s
25. F requency List 50 ms 10 s

< 2 GHz 0.6 dB +0.6 dB 0.19 dB

2 and 20 GHz 0.7 dB +0.7 dB 0.35 dB

> 20 and 26.5 GHz 0.9 dB +0.9 dB 0.38 dB

< 2 GHz 1.2 dB +1.2 dB 0.21 dB

2 and 20 GHz 1.3 dB +1.3 dB 0.46 dB

2 and 20 GHz 1.2 dB 0.25 dB

> 20 and 26.5 GHz 1.6 dB 0.27 dB

0.01 to 26.5 GHz +10.0 dBm 1.7 dB

0.01 to 20 GHz +13.0 dBm 1.7 dB

2-156 P erformance T ests

Agilent 83630L T est Record

Page 246 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 246 · click to enlarge
Page 246

T able 2-35. T est Record for 83630L (4 of 5)

Agilent Technologies 83640B — 08360125.pdf diagram (page 246)
Fig. 75 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertain t y

0.01 to 26.5 GHz +8.0 dBm 1.7 dB

0.01 to 20 GHz +11.5 dBm 1.7 dB

7. Minim um P ow er 0.2 mV 3 V
10. Maximum P ow er 0.5 V 20 V
8. 0.01 to < 2.0 GHz 30 dBc 1.75 dB
21. 2.0 and 26.5 GHz 50 dBc 1.75 dB
12. 7 and 20 GHz 50 dBc 1.75 dB
21. > 20 and 26.5 GHz 50 dBc 1.75 dB

100 MHz Fixed 60 dBc 1 dB

100 MHz O set 60 dBc 1 dB

LO F eedthrough 60 dBc 1 dB

20. Low Band Mixer Spur 60 dBc 1.85 dB

Agilent 8360 P erformance T ests

Agilent 83630L T est Record

Page 247 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 247 · click to enlarge
Page 247

T able 2-35. T est Record for 83630L (5 of 5)

Agilent Technologies 83640B — 08360125.pdf diagram (page 247)
Fig. 76 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertaint y

120 Hz Spur 55 dBc 1 dB

180 Hz Spur 55 dBc 1 dB

240 Hz Spur 55 dBc 1 dB

120 Hz Spur 55 dBc 1 dB

180 Hz Spur 55 dBc 1 dB

240 Hz Spur 55 dBc 1 dB

100 Hz to < 1 kHz 70 dBc 2 dB

1 kHz to < 10 kHz 78 dBc 2 dB

10 kHz to < 100 kHz 86 dBc 2 dB

100 kHz 107 dBc 2 dB

100 Hz to < 1 kHz 70 dBc 2 dB

1 kHz to < 10 kHz 78 dBc 2 dB

10 kHz to < 100 kHz 86 dBc 2 dB

100 kHz 107 dBc 2 dB

100 Hz to < 1 kHz 60 dBc 2 dB

1 kHz to < 10 kHz 68 dBc 2 dB

10 kHz to < 100 kHz 76 dBc 2 dB

100 kHz 97 dBc 2 dB

1 3 dB if the Invertron is not used.

2-158 P erformance T ests

Agilent 83630L T est Record

Page 248 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 248 · click to enlarge
Page 248

T able 2-36. T est Record for Agilent 83640L (1 of 5)

T est F acilityRep ort Num b er

Mo del Am bien t temp erature

Serial Num b er Relative humidit y %

Options Line frequency Hz (nominal)

Firm w are Revision

Sp ecial Notes:

Agilent 8360 P erformance T ests

Agilent 83640L T est Record

Show 4 more labels / short text on this page
Date · Customer · T ested by ·
Page 249 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 249 · click to enlarge
Page 249

T able 2-36. T est Record for 83640L (2 of 5)

Mo del Rep ort Num b er Date

T est Equipmen t Used Mo del Numb er T race Numb er Cal Due Date

1. Digitizing Oscilloscop e
2. Measuring Receiver
3. P ow er Sensor
4. P ow er Meter
5. Microw ave Sp ectrum Analyzer

2-160 P erformance T ests

Agilent 83640L T est Record

Page 250 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 250 · click to enlarge
Page 250

T able 2-36. T est Record for 83640L (3 of 5)

Agilent Technologies 83640B — 08360125.pdf diagram (page 250)
Fig. 77 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertain t y

10. Start F req 0.1% of 0.005% of
8. CW Step 50 ms 10 s
16. Stepp ed Sw eep 5.5 ms 1 s
25. F requency List 50 ms 10 s

< 2 GHz 0.6 dB 0.6 dB 0.22 dB

2 and 20 GHz 0.7 dB 0.7 dB 0.35 dB

> 20 and 40 GHz 0.9 dB 0.9 dB 0.38 dB

2 and 20 GHz 1.2 dB 0.25 dB

> 20 and 40 GHz 1.8 dB 0.27 dB

0.01 to 40 GHz +6.0 dBm 1.4 dB

0.01 to < 26.5 GHz +10.0 dBm 1.8 dB

0.01 to < 20 GHz +10.0 dBm 1.8 dB

Agilent 8360 P erformance T ests

Agilent 83640L T est Record

Page 251 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 251 · click to enlarge
Page 251

T able 2-36. T est Record for the 83640L (4 of 5)

Agilent Technologies 83640B — 08360125.pdf diagram (page 251)
Fig. 78 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertaint y

0.01 to 40 GHz +4 dBm 1.4 dB

0.01 to < 26.5 GHz +8 dBm 1.8 dB

0.01 to < 20 GHz +8.5 dBm 1.8 dB

7. Minim um P ow er 0.2 mV 3 V
10. Maximum P ow er 0.5 V 20 V
8. < 50 MHz 20 dBc 1.75 dB
8. 50 MHz and < 2.0 GHz 30 dBc 1.75 dB
21. 2.0 and < 26.5 GHz 50 dBc 1.75 dB
29. 26.5 to 40 GHz 40 dBc 1.75 dB
12. 7 and 20 GHz 50 dBc 1.75 dB
29. > 20 and 40 GHz 40 dBc 1.55 dB

100 MHz Fixed 60 dBc 1 dB

100 MHz O set 60 dBc 1 dB

LO F eedthrough 60 dBc 1 dB

2-162 P erformance T ests

Agilent 83640L T est Record

Page 252 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 252 · click to enlarge
Page 252

T able 2-36. T est Record for the 83640L (5 of 5)

Agilent Technologies 83640B — 08360125.pdf diagram (page 252)
Fig. 79 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertain t y

20. Low Band Mixer Spur 60 dBc 1.85 dB

100 Hz to < 1 kHz 70 dBc 2 dB

1 kHz to < 10 kHz 78 dBc 2 dB

10 kHz to < 100 kHz 86 dBc 2 dB

100 kHz 107 dBc 2 dB

100 Hz to < 1 kHz 70 dBc 2 dB

1 kHz to < 10 kHz 78 dBc 2 dB

10 kHz to < 100 kHz 86 dBc 2 dB

100 kHz 107 dBc 2 dB

100 Hz to < 1 kHz 60 dBc 2 dB

1 kHz to < 10 kHz 68 dBc 2 dB

10 kHz to < 100 kHz 76 dBc 2 dB

100 kHz 97 dBc 2 dB

1 3 dB if the Invertron is not used.

Agilent 8360 P erformance T ests

Agilent 83640L T est Record

Page 253 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 253 · click to enlarge
Page 253

T able 2-37. T est Record for Agilent 83650L (1 of 6)

T est F acilityRep ort Num b er

Mo del Am bien t temp erature

Serial Num b er Relative humidit y %

Options Line frequency Hz (nominal)

Firm w are Revision

Sp ecial Notes:

2-164 P erformance T ests

Agilnet 83650L T est Record

Show 4 more labels / short text on this page
Date · Customer · T ested by ·
Page 254 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 254 · click to enlarge
Page 254

T able 2-37. T est Record for 83650L (2 of 6)

Mo del Rep ort Num b er Date

T est Equipmen t Used Mo del Numb er T race Numb er Cal Due Date

1. Digitizing Oscilloscop e
2. Measuring Receiver
3. P ow er Sensor
4. P ow er Meter
5. Microw ave Sp ectrum Analyzer

Agilent 8360 P erformance T ests

Agilnet 83650L T est Record

Page 255 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 255 · click to enlarge
Page 255

T able 2-37. T est Record for 83650L (3 of 6)

Agilent Technologies 83640B — 08360125.pdf diagram (page 255)
Fig. 80 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertaint y

10. Start F req 0.1% of 0.005% of
8. CW Step: 50 ms 10 s
16. Stepp ed Sw eep 5.5 ms 1 s
25. F requency List 50 ms 10 s

< 2 GHz 0.6 dB 0.6 dB 0.22 dB

2 and 20 GHz 0.7 dB 0.7 dB 0.35 dB

> 20 and 40 GHz 0.9 dB 0.9 dB 0.38 dB

> 40 GHz 1.7 dB 1.7 dB 0.79 dB

2 and 20 GHz 1.2 dB 0.25 dB

> 20 and 40 GHz 1.6 dB 0.38 dB

2-166 P erformance T ests

Agilnet 83650L T est Record

Page 256 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 256 · click to enlarge
Page 256

T able 2-37. T est Record for 83650L (4 of 6)

Agilent Technologies 83640B — 08360125.pdf diagram (page 256)
Fig. 81 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertain t y

0.01 to 50 GHz +2.5 dBm 2.3 dB

0.01 to < 40 GHz +5.0 dBm 1.3 dB

0.01 to < 26.5 GHz +10.0 dBm 1.7 dB

0.01 to < 20 GHz +10.0 dBm 1.7 dB

0.01 to 50 GHz 0 dBm 2.3 dB

0.01 to < 40 GHz +3.0 dBm 1.3 dB

0.01 to < 26.5 GHz +8.0 dBm 1.7 dB

0.01 to < 20 GHz +8.5 dBm 1.7 dB

7. Minim um P ow er 0.2 mV 3 V
10. Maximum P ow er 0.5 V 20 V
8. < 50 MHz 20 dBc 1.75 dB
8. 50 MHz to < 2.0 GHz 30 dBc 1.75 dB
21. 2.0 to 26.5 GHz 50 dBc 1.75 dB
37. 26.5 to 50 GHz 40 dBc 2.22 dB
12. 7 to 20 GHz 50 dBc 1.75 dB
29. > 20 to 40 GHz 40 dBc 1.55 dB
37. > 40 to 50 GHz 35 dBc 2.22 dB

Agilent 8360 P erformance T ests

Agilnet 83650L T est Record

Page 257 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 257 · click to enlarge
Page 257

T able 2-37. T est Record for 83650L (5 of 6)

Agilent Technologies 83640B — 08360125.pdf diagram (page 257)
Fig. 82 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertaint y

100 MHz Fixed 60 dBc 1 dB

100 MHz O set 60 dBc 1 dB

LO F eedthrough 60 dBc 1 dB

20. Low Band Mixer Spur 60 dBc 1.85 dB

120 Hz Spur 55 dBc 1 dB

180 Hz Spur 55 dBc 1 dB

240 Hz Spur 55 dBc 1 dB

120 Hz Spur 55 dBc 1 dB

180 Hz Spur 55 dBc 1 dB

240 Hz Spur 55 dBc 1 dB

1 3 dB if the Invertron is not used.

2-168 P erformance T ests

Agilnet 83650L T est Record

Page 258 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 258 · click to enlarge
Page 258

T able 2-37. T est Record for 83650L (6 of 6)

Agilent Technologies 83640B — 08360125.pdf diagram (page 258)
Fig. 83 · 08360125.pdf

No. T est Description Sp ec. Results Sp ec. Uncertaint y

100 Hz to < 1 kHz 70 dBc 2 dB

1 kHz to < 10 kHz 78 dBc 2 dB

10 kHz to < 100 kHz 86 dBc 2 dB

100 kHz 107 dBc 2 dB

100 Hz to < 1 kHz 70 dBc 2 dB

1 kHz to < 10 kHz 78 dBc 2 dB

10 kHz to < 100 kHz 86 dBc 2 dB

100 kHz 107 dBc 2 dB

100 Hz to < 1 kHz 60 dBc 2 dB

1 kHz to < 10 kHz 68 dBc 2 dB

10 kHz to < 100 kHz 76 dBc 2 dB

100 kHz 97 dBc 2 dB

Agilent 8360 P erformance T ests

Agilnet 83650L T est Record

Page 259 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 259 · click to enlarge
Page 259
Page 259 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 260 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNN
Page 260 · click to enlarge
Page 260

How T o Use This Chapter

P erform these adjustmen ts only if directed b y the A gilent T e chnolo gies 8360 B-Series

Swept Signal Gener ator/ 8360 L-Series Swept CW Gener ator T r oublesho oting Guide or if a

p erformance test fails.

F old-out men u maps are lo cated in Chapter 6.

NNNNNNNNNNNNNN

Some men us ha v e more than one page of softkeys. Select the

More softkey to view the next

page of softkeys.

NNNNNNNNNNNNNN

More is not included in the k eystrok es given in these pro cedures.

In all cases where y ou are instructed to preset the syn thesizer, use the factory

preset mo de only .

If a password is set, it will ha v e to b e disabled b efore the adjustmen t men u

can b e accessed. See Chapter 4 for instructions.

The equipmen t required to p erform the adjustmen ts in this c hapter is listed in Chapter 1.

Y ou may use an y equipmen t that satis es the critical sp eci cations given. Use a non-metal lic

adjustmen t to ol.

This instrumen t has b een designed in accordance with in ternational safety standards.

Ho w ev er, for safe op eration follow the cautions and w arnings in this man ual.

Agilent 8360 Adjustments 3-1

Adjustments · Menu Maps · Equipment Required

Page 261 of the Agilent Technologies 83640B user manual — CA UTION
Page 261 · click to enlarge
Page 261

W ARNING

These servicing instructions are for use b y qualified personnel only . T o a v oid

electrical shock, do not perform any servicing unless y ou are qualified to do

so.

The opening of co v ers or remo v al of parts is lik ely to expose dangerous

v oltages. Disconnect the instrument from all v oltage sources while it is being

opened.

The detachable po w er cord is the instrument disconnecting device. It

disconnects the mains circuits from the mains supply before other parts of

the instrument. The front panel switch is only a standb y switch and is not a

LINE switch.

The po w er cord is connected to internal capacitors that may remain liv e for 5

seconds after disconnecting the plug from its po w er supply .

This is a Safety Class I product (pro vided with a protectiv e earthing ground

incorporated in the po w er cord). The mains plug shall only be inserted in a

sock et outlet pro vided with a protectiv e earth contact. Any interruption of the

protectiv e conductor, inside or outside the instrument, is lik ely to mak e the

instrument dangerous. Intentional interruption is prohibited.

For continued protection against fire hazard replace line fuse only with same

type and rating (F 5A/250 V). The use of other fuses or material is prohibited.

CA UTION

Always use the three-prong ac p ow er cord supplied with this instrumen t.

F ailure to ensure adequate earth grounding b y not using this cord may cause

instrumen t damage.

Before switching on this pro duct, make sure that the line v oltage selector

switch is set to the v oltage of the p ow er supply and the correct fuse is

installed. Assure the supply v oltage is in the sp eci ed range.

3-2 Adjustments Agilent 8360

Page 262 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 262 · click to enlarge
Page 262

The follo wing is a list of this c hapter's adjustmen ts and the page n um b er that eac h

adjustmen t is lo cated on:

1. 10 MHz Standard : 3-6
2. Sw eep Ramp : 3-12
3. Sampler Assem bly : 3-13
4. YO Driv er +10 V Reference : 3-18
5. YO Driv er Gain and Linearity : 3-20
6. YO Lo op Gain : 3-23
7. Ampli er/Multiplier Adjustmen ts : 3-25
8. Ampli er/Filter Adjustmen ts : 3-43
9. Lo w P ow er SRD Bias : 3-79
10. Mo dulator O set and Gain : 3-81
11. ALC P ow er Lev el Accuracy : 3-86
12. P ow er Flatness : 3-92

(this adjustmen t may b e replaced with the automated

p ow er atness adjustmen t in Chapter 5.)

13. AM/FM DA C O set and Gain (Option 002) y : 3-95
14. AM Accuracy y : 3-97
15. AM O set (Agilent 8360 L-Series Only) : 3-101
16. AM Delay y : 3-103
17. FM Gain y : 3-106
18. Square W av e Symmetry : 3-109
19. AM Input Imp edance y : 3-111
20. FM Input Imp edance y : 3-114
21. Mo dulation Generator Flatness (Option 002) y : 3-117
22. External T rigger Edge Selection : 3-120
23. V olts/GHz : 3-122
24. A9 Pulse Board Jump er (W1) : 3-124
25. Selftest P atches (8360 L-Series Only) : 3-125

Adjustments for 8360 B-Series Only

The adjustmen ts that apply to the

8360 B-series swept signal generators

only are marked with the following symb ol: y

F or the following automated adjustmen ts, refer to Chapter 5, \Automated T ests":

YO Delay Adjustmen t

P ow er Flatness

Agilent 8360 Adjustments 3-3

List of Adjustments · ADC Adjustmen t

Page 263 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 263 · click to enlarge
Page 263
1. 10 MHz Standard

This pro cedure adjusts the frequency accuracy of the in ternal 10 MHz time base. This

adjustmen t should b e done on a regular basis if absolute frequency accuracy is imp ortant (see

Figure 3-3.)

F or b est accuracy , readjust the 10 MHz timebase oscillator after the syn thesizer has b een

on or in standb y for 24 hours. See \Accuracy V ersus Adjustmen t In terv al," following this

adjustmen t, for information on ho w to determine a p erio dic adjustmen t sc hedule.

After the timebase is adjusted, the timebase frequency should stay within the aging rate if the

follo wing things happ en:

The time base ov en do es not co ol do wn.

The instrumen t k eeps the same orientation with resp ect to the earth's magnetic eld.

The instrumen t stays at the same altitude.

The instrumen t do es not receiv e an y mec hanical sho c k.

If the time base ov en co ols (the instrumen t is disconnected from A C p ow er), y ou may ha v e to

readjust the time base frequency after a new w arm-up cycle. T ypically , ho w ev er, the time base

frequency returns to within 1 Hz of the original frequency .

Y ou can adjust the in ternal timebase after reconnecting A C p ow er for 10

minutes, but for b est accuracy , test again after the instrumen t has b een on or

in standb y for 24 hours.

F requency c hanges, due either to a c hange in orientation with resp ect to the earth's magnetic

eld or to a c hange in altitude, are usually eliminated when the instrumen t is returned to its

original p osition. A frequency c hange due to mec hanical sho c k usually app ears as a xed

frequency error.

Description and Procedure · 3-4 Adjustments · 10 MHz Standard

Page 264 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 264 · click to enlarge
Page 264
1. Connect the equipmen t as sho wn in Figure 3-1 . Preset all instrumen ts and let them w arm

up for one hour.

Agilent Technologies 83640B — 08360125.pdf diagram (page 264)
Fig. 84 · 08360125.pdf

Figure 3-1. 10 MHz Standard Adjustment Setup

Digitizing Oscilloscop e HP/Agilent 54111D

F requency Standard HP/Agilent 5061A

If the oscilloscop e do es not ha v e a 50 input imp edance, connect c hannel 1

through a 50 feedthrough.

2. On the oscilloscop e, set:

Channel 1:

Channel 2:

Timebase:

Trigger:

Trigger Source Trig 3

Display:

Display Mode Real Time

3. On the oscilloscop e, adjust the trigger lev el so that the sweep is sync hronized to the

syn thesizer's in ternal standard. The w av eform will app ear to drift.

Show 12 more labels / short text on this page
Equipment · Display On · Volts/Division 200 mV · Input Coupling dc · Input Impedance 50 · Display Off · Time/Division 10 ns · Trigger External · Trigger Mode Edge · Input Coupling ac · Agilent 8360 Adjustments · 10 MHz Standard
Page 265 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 265 · click to enlarge
Page 265
4. Using a non-metallic to ol, adjust the A23 10 MHz standard (see Figure 3-2) for minimum

horizontal mov emen t of the oscilloscop e w av eform.

Agilent Technologies 83640B — 08360125.pdf diagram (page 265)
Fig. 85 · 08360125.pdf

Figure 3-2. 10 MHz Standard Adjustment Location

3-6 Adjustments · 10 MHz Standard

Page 266 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 266 · click to enlarge
Page 266

Related P erformance T ests

In ternal TimebaseAging Rate

In Case of Difficulty

1. Ensure that an external standard is not connected. (At instrumen t preset the syn thesizer

automatically c ho oses the external standard as the reference if one is connected to the 10

MHz REF INPUT.)

2. Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

Agilent 8360 Adjustments · 10 MHz Standard

Page 267 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 267 · click to enlarge
Page 267

Accuracy V ersus Adjustment Interv al

Figure 3-3 sho ws the required adjustmen t in terv al to maintain a given accuracy . If y ou kno w

the aging rate, y ou can determine a more precise adjustmen t in terv al.

Agilent Technologies 83640B — 08360125.pdf diagram (page 267)
Fig. 86 · 08360125.pdf

Figure 3-3. Accuracy V ersus Adjustment Interv al

1. Find the line on Figure 3-3 that corresp onds to the 10 MHz timebase oscillator aging rate.

(T o determine the aging rate, see \Performance T ests.")

2. On the horizontal axis, nd the maximum allo w able error (accuracy) that y ou w an t.
3. F ollo w the maximum allo w able error v ertically un til it in tersects the kno wn aging rate.
4. F rom the p oint of in tersection, mov e horizontally to the left and read the minimum

calibration in terv al measured in da ys.

Y ou can also determine the minimum calibration in terv al using the following formula:

maximum allo w able error

= calibration in terv al in da ys

kno wn aging rate (p er da y)

An Alternate Method · 3-8 Adjustments · 10 MHz Standard

Page 268 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 268 · click to enlarge
Page 268
2. Sw eep Ramp

No test equipmen t is required for this adjustmen t.

This is an automatic adjustmen t. The syn thesizer's in ternal DVM measures the sweep

ramp v oltage at the end of the sweep. The syn thesizer then sets the sweep time calibration

constants to calibrate the sweep ramp.

1. T urn the syn thesizer on and press

PRESET

. Let the syn thesizer w arm up for at least one

hour.

2. Select the A14 sweep ramp adjustmen t. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SERVICE

The syn thesizer will measure, calculate, and store the sweep time calibration constants.

When the adjustmen t is complete, the follo wing message app ears on the display:

Sweep Ramp Cal Completed

Calibration constants were modified.

3. If this is the last calibration constant y ou will b e adjusting, see \Calibration Constants" to

store the calibration constants as protected data

(in EEPR OM).

Related P erformance T ests

Sw ept F requency Accuracy

In Case of Difficulty

Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

Show 9 more labels / short text on this page
Description and Procedure · · · Adjust Menu · AssyAdj Menu · A14 SwpRmp · Self-T ests · Agilent 8360 Adjustments · Sw eep Ramp
Page 269 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 269 · click to enlarge
Page 269
3. Sampler Assembly

No test equipmen t is required for this pro cedure.

There are three sampler assembly adjustmen ts:

In the sampler match adjustmen t, as the in ternal DVM monitors the lo op input to the

sampler, trim capacitors are adjusted to optimize the sampler match ov er the full frequency

range of the lo op.

The 200 MHz lo op gain adjustmen t is automatic. When this adjustmen t is selected in the

A6 adjustmen t men u, the syn thesizer selects in ternal calibration constants for optimum lo op

gain ov er the full frequency range of the lo op.

In the IF gain adjustmen t, the IF ampli er gain is adjusted as the in ternal DVM monitors

the IF output lev el to the YO lo op.

1. T urn the syn thesizer to standb y . A t FL2 (see Figure 3-4), disconnect the semi-rigid cable

b et w een FL2 and the A6 assembly .

Agilent Technologies 83640B — 08360125.pdf diagram (page 269)
Fig. 87 · 08360125.pdf

Figure 3-4. FL2 Location

2. Lea ving the exible cable connected to A6J1 (40 MHz reference), place the A6 assembly

on the analog extender b oard.

3. T urn the syn thesizer on and press

. Let it w arm up for at least one hour.

PRESET

4. Select the A6 adjustmen t men u. Press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SERVICE

Show 8 more labels / short text on this page
Description and Procedure · · · Adjust Menu · AssyAdj Menu · A6 Adj Menu · 3-10 Adjustments · Sampler Assembly
Page 270 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNN
Page 270 · click to enlarge
Page 270

200 MHz Loop V CO Tune Check

In the A6 adjustmen t men u, select the A6 V CO tune c hec k

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

A6 VCO Tune .

6. The follo wing message should b e display ed:

Ctrim and C60 are OK.

If this message is not display ed, refer to troublesho oting.

If the three tuning v oltages at the b ottom of the display (V200, V avg, and

V220) are all equal, the 200 MHz phase-lo c k ed lo op is unlo c k ed.

NNNNNNNNNNNNNN

7. Select

Done .

In the A6 adjustmen t men u, select the A6 sampler match adjustmen t,

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

A6 Smplr Match .

The sampler match display is a graphic represen tation of the sampler match

ov er the full frequency range of the 200 MHz lo op. The more asterisks

display ed at a frequency , the more output signal (and the b etter the match) at

that frequency . Make the adjustmen t for b est match ov er the full frequency

range of the lo op.

9. Adjust A6C82 and A6C83 ( Figure 3-5 ) for b est ov erall sampler match.
Agilent Technologies 83640B — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 270)
Fig. 88

Figure 3-5. Sampler Match Adjustment Locations

NNNNNNNNNNNNNN

Done .

Show 4 more labels / short text on this page
Sampler Match Adjustment · Select · Agilent 8360 Adjustments · Sampler Assembly
Page 271 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNN
Page 271 · click to enlarge
Page 271

200 MHz Loop Gain Adjustment

In the A6 adjustmen t men u, select the A6 lo op gain adjustmen t,

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

A6 Loop Gain . The

syn thesizer automatically p erforms the adjustmen t. When the message Calibration

Constants should be SAVED is display ed, the adjustmen t is complete. Con tinue with the

next step.

12. Set the syn thesizer to standb y . Reinstall the A6 assembly in the instrumen t, and connect

all cables.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

T urn the syn thesizer on. Select

A6 IF Gain .

14. Adjust A6R73 (see Figure 3-6) to cen ter the \needle" on the display .
Agilent Technologies 83640B — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 271)
Fig. 89

Figure 3-6. IF Gain Adjustment Location

NNNNNNNNNNNNNN

Done .

16. If this is the last calibration constant y ou will b e adjusting, see \Calibration Constants"

to store the calibration constants as protected data (in EEPR OM).

Related P erformance T ests

In Case of Difficulty

1. Make sure the 40 MHz reference input cable is connected while the A6 assembly is on the

analog extender b oard.

2. F or the IF gain adjustmen t, if the needle registers on the far left or right, ensure that all

coax cables (including semi-rigid) are connected and that no unlo c k ed message is display ed.

Show 5 more labels / short text on this page
IF Gain Adjustment · Select · Self-tests · 3-12 Adjustments · Sampler Assembly
Page 272 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 272 · click to enlarge
Page 272
3. There is some in teraction b et w een the adjustmen ts in this pro cedure. T ry an alternate

order of adjustmen ts.

4. Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

Agilent 8360 Adjustments · Sampler Assembly

Page 273 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNN
Page 273 · click to enlarge
Page 273
4. YO Driv er +10 V Reference

No test equipmen t is required for this pro cedure.

As the in ternal DVM monitors the +10 V reference, y ou make the adjustmen t.

1. T urn the syn thesizer on and press

. Let the instrumen t w arm up for at least one

PRESET

hour.

2. Select the A13 adjustmen t men u. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SERVICE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Select the A13 +10 V reference adjustmen t,

A13+10V Ref .

4. Adjust A13R11 (see Figure 3-7) to cen ter the \needle" on the display .
Agilent Technologies 83640B — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 273)
Fig. 90

Figure 3-7. YO Driv er +10 V Reference Adjustment Location

NNNNNNNNNNNNNN

Done .

Related P erformance T ests

In Case of Difficulty

Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

YO Driv er +10 V Reference

Show 9 more labels / short text on this page
Description and Procedure · · · Adjust Menu · AssyAdj Menu · A13 Adj Menu · Select · Self-tests · 3-14 Adjustments
Page 274 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNN
Page 274 · click to enlarge
Page 274
5. YO Driv er Gain and Linearity

No test equipmen t is required for these adjustmen ts.

F or the gain adjustmen t, the syn thesizer's in ternal DVM monitors the YO lo op error v oltage

at tw o frequencies (p oints B and C in Figure 3-8). When y ou select the gain p oten tiometer

adjustmen t, the syn thesizer displays the di erence in v oltage b et w een p oints B and C as a

\needle" with arrows on either side. By adjusting A13R54 to cen ter the \needle", y ou equalize

the v oltage b et w een the tw o frequency p oints.

The linearit y adjustmen t has tw o parts (eac h of whic h is selected and adjusted similarly to the

gain adjustmen t). The low break p oint adjustmen t equalizes the v oltage b et w een p oints A and

B in Figure 3-8. The high break p oint adjustmen t equalizes the v oltage b et w een p oints C and

D.

Y ou m ust adjust the gain rst; then adjust the break p oints.

Agilent Technologies 83640B — 08360125.pdf diagram (page 274)
Fig. 91 · 08360125.pdf

Figure 3-8. YO Gain and Linearity Break P oints

1. T urn the syn thesizer on and press

. Let it w arm up for at least one hour.

PRESET

2. Select the A13 gain men u. Press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SERVICE

If the follo wing message is display ed: Adjust A13 pot until "locked".,

adjust the p oten tiometer un til the lo c k ed indication is display ed. Con tin ue

with step 3.

3. Select the gain p oten tiometer adjustmen t,

NNNNNNNNNNNNNNNNNNNNNNNNNN

Gain Pot .

4. Adjust A13R54 (see Figure 3-9 ) to cen ter the \needle" on the display .

YO Driv er Gain and Linearity

Show 8 more labels / short text on this page
Description and Procedure · · · Adjust Menu · AssyAdj Menu · A13 Adj Menu · A13 Gain Menu · Agilent 8360 Adjustments
Page 275 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNN
Page 275 · click to enlarge
Page 275
Agilent Technologies 83640B — 08360125.pdf diagram (page 275)
Fig. 92 · 08360125.pdf

Figure 3-9. YO Gain and Linearity Adjustment Locations

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Select the low break p oint p oten tiometer adjustmen t,

Lo Bk Pot .

6. Adjust A13R90 (see Figure 3-9 ) to cen ter the \needle" on the display .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Select the high break p oint p oten tiometer adjustmen t,

Hi Bk Pot .

8. Adjust A13R92 (see Figure 3-9 ) to cen ter the \needle" on the display .

NNNNNNNNNNNNNN

Done .

10. If this is the last calibration constant y ou will b e adjusting, see \Calibration Constants"

to store the calibration constants as protected data (in EEPR OM).

Related P erformance T ests

In Case of Difficulty

Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

YO Driv er Gain and Linearity

Select · Self-tests · 3-16 Adjustments

Page 276 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNN
Page 276 · click to enlarge
Page 276
6. YO Loop Gain

No test equipmen t is required for this pro cedure.

1. T urn the syn thesizer on and press

PRESET

. Let the instrumen t w arm up for at least one

hour.

2. Select the assembly adjustmen t men u. Press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SERVICE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Select the A5 lo op gain adjustmen t,

A5 Loop Gain .

4. On the A5 assembly , set all v e switches on A5S1 to the closed p osition. (See Figure 3-10.)
Agilent Technologies 83640B — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 276)
Fig. 93

Figure 3-10. Switch and Adjustment Locations

5. Adjust A53R80 (see Figure 3-10 ) to cen ter the \needle" on the display .

NNNNNNNNNNNNNN

Done .

NNNNNNNNNNNNNN

Set all A5 switches to op en and select

Done .

Related P erformance T ests

In Case of Difficulty

1. Be sure to close the A5S1 switches b efore making the adjustmen t and op en the switches

when the adjustmen t is complete.

2. Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

Show 9 more labels / short text on this page
Description and Procedure · · · Adjust Menu · AssyAdj Menu · Select · None · Agilent 8360 Adjustments · YO Loop Gain
Page 277 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 277 · click to enlarge
Page 277
7. Amplifier/Multiplier Adjustments

This pro cedure maximizes RF p ow er b y tracking the tuned lter in the ampli er/multiplier

to the RF output frequency . The ampli er/ lter output is also monitored. Initial tracking

is done in single band sweeps at slow sweep sp eeds to eliminate the e ects of bandcross and

h ysteresis in the tuning coil. (Those will b e corrected with dela y and risetime calibration

constants.) A squegging clamp adjustmen t limits the p ow er in to the ampli er/m ultiplier

and is adjusted in the ampli er/ lter adjustmen ts for maximum output p ow er without

squegging. A \squegg" is a distortion in the RF output caused b y to o m uc h p ow er to the

ampli er/multiplier and c haracterized b y a p ow er drop out in a p ortion of the trace, or

a p ow er drop out ov er a broad frequency range. The SRD bias adjustmen ts are made to

optimize the eciency of the frequency m ultiplication. The YTM dela y comp ensation

adjustmen ts maximize p ow er for fast single- and m ulti-band sweeps.

If y ou ha v e replaced either the YO, the YO driver, or the m ultiplier/ lter

driver, rst initiate an auto tracking (press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USER CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Auto Track .) If auto tracking passes and the instrumen t passes the

\Maximum Lev eled P ow er" p erformance test, do not con tinue with this

pro cedure. If auto tracking fails, (an error message is display ed), con tinue with

this pro cedure.

Show 5 more labels / short text on this page
Description · · Tracking Menu · 3-18 Adjustments · Amplifier/Multiplier Adjustments
Page 278 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 278 · click to enlarge
Page 278

The follo wing explanations are provided for a b etter understanding of the ampli er/multiplier

adjustmen ts. Refer to these explanations as often as necessary . They will help make these

adjustmen ts easier. (The adjustmen t pro cedure follo ws.)

Offset and Gain Adjustments.

F or all o set and gain adjustmen ts, adjust thr ough the bandpass. Keep adjusting un til the

p ow er p eaks and then drops o . Then reset the adjustmen t to the p eak ed p oint.

F or all o set and gain adjustmen ts, if p ow er stays p eak ed ov er sev eral calibration constant

v alues, set the calibration constant to the middle v alue of the p eak ed range.

Adjusting to the Center of the P assband.

The ampli er/multiplier has a tracking lter on its output. Its purp ose is to pass the

selected RF output frequency and reject harmonics. The ob jective of this pro cedure is to

track the bandpass lter with the RF output frequency ov er a v ariety of sweep conditions

and frequency ranges. E ects of the auxiliary output may mask the true cen ter of the

bandpass. Always tune to the low er p eak (lo w er calibration constant v alue).

SRD Bias Adjustment.

Adjust these calibration constants into sque gging and then bac k o un til p eak p ow er is

ac hiev ed.

After reac hing p eak p ow er, decrease the calibration constant v alue b y 15 coun ts when

adjusting the A term and decrease p ow er b y 1/2 dB (approximately 1/3 division) when

adjusting the B term.

Single Band Delay Compensation Adjustments.

If a drop in p ow er greater than 1 division o ccurs when in single sweep, increase the A

dela y term to put the A term at the high end of the p eak. F or example, for the following

calibration constant v alues:

Start of the p eak = 1800

Middle of the p eak = 1950

End of the p eak = 2100

Set the A term to 2050

Setting the F astest Sw eep Time.

Setting the sweep time to 0 milliseconds sets the syn thesizer to the fastest sweep time for

the frequency range b eing swept. Eac h time the frequency range is c hanged, the sweep time

m ust b e reset to 0 ms in order to maintain the fastest sweep time.

Sw eep Speed Related Adjustments.

Delay comp ensation adjustmen ts are a ected b y sweep sp eed. The adjustmen ts are

p erformed at fast sweep sp eeds whic h are the w orst case.

All other adjustmen ts are p erformed at slow sweep sp eeds; they are not a ected b y sweep

sp eed.

If y ou ha v e a p ow er problem that o ccurs only for fast sweep rates, it is probably a ected b y

dela y comp ensation or risetime adjustmen ts.

Adjustment Help · Agilent 8360 Adjustments · Amplifier/Multiplier Adjustments

Page 279 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 279 · click to enlarge
Page 279

YTM Bandcross Delay T erms.

The YTM Bx Dly terms are either o set or gain terms used only during m ulti-band sweeps.

A1 denotes the o set term for band 1. B1 denotes the gain term for band 1. The o set

(A) should b e adjusted to maximize p ow er at the b eginning of the band. The gain (B) is

adjusted to maximize p ow er tow ard the end of the band.

General T echniques.

In some adjustmen ts p ow er may p eak twice. This is caused b y the auxiliary output masking

the true p eak. Always adjust to the rst p eak (low est v alue of the calibration constant).

If the adjustmen t has a wide range without an y apparen t improv emen t. Lea v e it set to the

default v alue.

3-20 Adjustments · Amplifier/Multiplier Adjustments

Page 280 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 280 · click to enlarge
Page 280

In this pro cedure, some calibration constants and p oten tiometers may b e

set to default conditions and left. This is normal. These b oards con tain

adjustmen ts for di eren t vin tages of the 8360 family . Not all adjustmen ts are

used eac h time.

1. Connect the equipmen t as sho wn in Figure 3-11 using an analog oscilloscop e with A

v ersus B sweep capabilit y . P ow er on all the instrumen ts and let them w arm up for at least

one hour.

2. On the syn thesizer, press

PRESET

.

Agilent Technologies 83640B — PRESET diagram (page 280)
Fig. 94

Figure 3-11. Amplifier/Multiplier Adjustments Setup

Analog Oscilloscop e HP/Agilent 1740A

Oscilloscop e Prob es HP/Agilent 10431A

Detector HP/Agilent 33330D (Agilent 83620B /22B/24B,

Agilent 83623B/L, and Agilent 83630B/L)

Detector HP/Agilent 33330E (Agilent 83640B/L & Agilent 83650B/L)

A tten uator 10 dB HP/Agilent 8493C Option 010 (83620B /22B/24B,

A tten uator 10 dB HP/Agilent 8490D Option 010

3. Set the default v alues given in T able 3-1 in to the corresp onding YTM calibration

constants. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SERVICE

Show 10 more labels / short text on this page
Procedure · Equipment · 83623B/L, and 83630B/L) · (83640B/L and 83650B/L) · · · Adjust Menu · Calib Menu · Agilent 8360 Adjustments · Amplifier/Multiplier Adjustments
Page 281 of the Agilent Technologies 83640B user manual — ENTER
Page 281 · click to enlarge
Page 281

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Select Cal En ter the n um b er of the rst calibration constant from T able 3-1 and

terminate the en try with

ENTER

. When using the front panel em ulator softw are, pressing

the arrow k eys will allo w y ou to v ary the calibration constants. The calibration constant

sho wn at the top of the \display" window is the active constant.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Modify Cal En ter the default v alue from T able 3-1 and terminate the en try with

.

ENTER

Using the up/do wn arrow k eys, select eac h of the rest of the calibration constants in

T able 3-1 and set their default v alues b y en tering the v alue on the n umeric k eypad.

T erminate y our en try with the

ENTER

k ey .

T able 3-1. Calibration Constant Descriptions

Numb er Description Default

105 SRD Bias A; Band 1 255 Maximize p ow er over low er 30%

106 SRD Bias A; Band 2 130

of band. Sw eep sp eed is not a

factor. Adjust for 0.5 dB

107 SRD Bias A; Band 3 130

b elow maxim um p ow er.

108 SRD Bias A; Band 4 130

109 SRD Bias A; Band 5 130

110 SRD Bias A; Band 6 120

111 SRD Bias A; Band 7 120

118 SRD Bias B; Band 1 255 Maximize p ow er over higher 30%

119 SRD Bias B; Band 2 150

of band. Sw eep sp eed is not a

factor. Adjust for 0.5 dB b elow

120 SRD Bias B; Band 3 150

maxim um p ow er.

121 SRD Bias B; Band 4 150

122 SRD Bias B; Band 5 160

123 SRD Bias B; Band 6 160

124 SRD Bias B; Band 7 160

131 Squegg Clamp 1A 255 Default v alues are not altered.

132 Squegg Clamp 2A 255

Num b ers should remain

unchanged (#131 is adjusted

133 Squegg Clamp 3A 255

in another pro cedure).

134 Squegg Clamp 4A 255

135 Squegg Clamp 5A 255

136 Squegg Clamp 6A 255

137 Squegg Clamp 7A 255

144 YTM Delay T erm A Hrm 1 1000 Default v alues are not altered.

Num b ers should remain

145 YTM Delay T erm A Hrm 2 1000

unchanged. Maximize p ow er over

146 YTM Delay T erm A Hrm 3 1000

rst 20% of band. A ects

fast sw eeps only . Single sw eep

147 YTM Delay T erm A Hrm 4 1000

mo de is also critical.

148 YTM Delay T erm A Hrm 5 1000

149 YTM Delay T erm A Hrm 6 1000

150 YTM Delay T erm A Hrm 7 1000

Show 5 more labels / short text on this page
Adjustment · V alue · Description · 3-22 Adjustments · Amplifier/Multiplier Adjustments
Page 282 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 282 · click to enlarge
Page 282

T able 3-1. Calibration Constant Descriptions (continued)

Numb er Description Default

157 YTM Delay T erm B Hrm 1 300 Default v alues are not altered.

Num b ers should remain

158 YTM Delay T erm B Hrm 2 300

unchanged. Maximize p ow er

159 YTM Delay T erm B Hrm 3 300

over higher 30% of band.

160 YTM Delay T erm B Hrm 4 300

A ects fast sw eeps only .

161 YTM Delay T erm B Hrm 5 300

162 YTM Delay T erm B Hrm 6 300

163 YTM Delay T erm B Hrm 7 300

170 YTM Bx Dly T erm A1 1000 Default v alues for all

terms except A3 and A4.

171 YTM Bx Dly T erm A2 1000

Delay comp ensation for

172 YTM Bx Dly T erm A3 1000

m ulti-band sw eeps only .

173 YTM Bx Dly T erm A4 1000

Adjust for maxim um p ow er at

low er 50% of band. A ects

174 YTM Bx Dly T erm A5 1000

fast sw eep sp eeds only .

175 YTM Bx Dly T erm A6 1000

176 YTM Bx Dly T erm A7 1000

183 YTM Bx Dly T erm B1 300 Default v alues for all

terms except B3 and B4.

184 YTM Bx Dly T erm B2 300

Delay comp ensation for

185 YTM Bx Dly T erm B3 300

m ulti-band sw eeps only .

186 YTM Bx Dly T erm B4 300

Adjust for maxim um p ow er at

low er 50% of band. A ects

187 YTM Bx Dly T erm B5 300

fast sw eep sp eeds only .

188 YTM Bx Dly T erm B6 300

189 YTM Bx Dly T erm B7 300

195 YTM Kic k Threshold 110 Default v alues are not altered.

Num b ers should remain

196 YTM CW Kic k Max 600

unchanged.

197 YTM Mono Band Kic k 2000

198 YTM Stereo Band Kic k 2000

199 YTM Slew Rate 400

200 YTM Slew Max 50

201 YTM Slew Min 0

202 YTM Neg Kic k W ait 2

203 YTM Fwd Kic k Pct 50

204 YTM Fwd Kic k W ait 30

Show 5 more labels / short text on this page
Adjustment · V alue · Description · Agilent 8360 Adjustments · Amplifier/Multiplier Adjustments
Page 283 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 283 · click to enlarge
Page 283

T able 3-1. Calibration Constant Descriptions (continued)

Numb er Description Default

205 YTM Rise; Band 1 15 Minimize p ow er drop outs at the

start of band. Adjustmen t is only

206 YTM Rise; Band 2 A1 20

e ectiv e at fast sw eep sp eeds.

207 YTM Rise; Band 2 B1 2

Adjusted in the Ampli er/

208 YTM Rise; Band 2 A2 15

Filter Adjustments.

209 YTM Rise; Band 3 A1 70

210 YTM Rise; Band 3 B1 7

211 YTM Rise; Band 3 A2 15

212 YTM Rise; Band 3 B2 15

213 YTM Rise; Band 3 A3 15

214 YTM Rise; Band 4 15

215 YTM Rise; Band 5 15

216 YTM Rise; Band 6 10

217 YTM Rise; Band 7 10

225 *YTM Gain Band 1 2048 Maximize p ow er over higher 10%

226 *YTM Gain Band 2 2048

of band. Adjust at slow sw eep

sp eeds.

227 *YTM Gain Band 3 2048

228 *YTM Gain Band 4 2048

229 *YTM Gain Band 5 2048

230 *YTM Gain Band 6 2048

231 *YTM Gain Band 7 2048

238 *YTM O set Band 1 2048 Maximize p ow er over low er 10%

of band. Adjust at slow sw eep

239 *YTM O set Band 2 2048

sp eeds.

240 *YTM O set Band 3 2048

241 *YTM O set Band 4 2048

242 *YTM O set Band 5 2048

243 *YTM O set Band 6 2048

244 *YTM O set Band 7 2048

248 YTM B2 O set O set 0 Default v alues are not altered.

Num b ers should remain

unchanged.

Show 5 more labels / short text on this page
Adjustment · V alue · Description · 3-24 Adjustments · Amplifier/Multiplier Adjustments
Page 284 of the Agilent Technologies 83640B user manual — ALC
Page 284 · click to enlarge
Page 284

T able 3-1. Calibration Constant Descriptions (continued)

Numb er Description Default

373 YTM Dly T erm C Hrm 1 0 Default v alues are not altered.

Num b ers should remain

374 YTM Dly T erm C Hrm 2 0

unchanged.

375 YTM Dly T erm C Hrm 3 0

376 YTM Dly T erm C Hrm 4 0

377 YTM Dly T erm C Hrm 5 0

378 YTM Dly T erm C Hrm 6 0

379 YTM Dly T erm C Hrm 7 0

382 YTM Bx Dly T erm C1 0 Default v alues are not altered.

Num b ers should remain

383 YTM Bx Dly T erm C2 0

unchanged.

384 YTM Bx Dly T erm C3 0

385 YTM Bx Dly T erm C4 0

386 YTM Bx Dly T erm C5 0

387 YTM Bx Dly T erm C6 0

525 Hib ernation Time 90 Default v alues are not altered.

526 Bandcross 1 42

Num b ers should remain

unchanged.

527 Bandcross 2 32

528 Bandcross 3 28

529 Bandcross 4 26

530 Bandcross 5 28

531 Bandcross 6 20

532 Bandcross 7 20

Band 1 Adjustment.

4. On the syn thesizer, press:

ST ART

2.0

STOP

SWEEP TIME

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Leveling Point Ext Det (asterisk on)

ALC

POWER LEVEL

5. On the oscilloscop e, set:

Channel A:

Volts/Division5 mV/Division
OffsetAs required
Input CouplingDC
Input Impedance1 M
Volts/Division1 V/Division
OffsetAs required
Input CouplingDC
Input Impedance1 M
Show 28 more labels / short text on this page
Adjustment · V alue · Description · YTM Gain Adjustment · · · GHz · · · · 7 · GHz · · · · 200 · msec · · · · · · 25 · dBm · · Channel B · Agilent 8360 Adjustments · Amplifier/Multiplier Adjustments
Page 285 of the Agilent Technologies 83640B user manual — STOP
Page 285 · click to enlarge
Page 285
Sweep ModeA versus B

Adjust Channel B o set, v olts/division, and horizontal p osition con trols for a trace that

lls the full horizontal display . Adjust the v ertical p osition with the A v ertical p osition

con trol.

6. Make certain that calibration constant #131, Squegg Clamp 1A, is set to 255.

Then adjust A12R69, YTMG, to maximize high end p ow er (last tw o horizontal display

divisions).

7. Mo dify calibration constant #238, YTM O set; Band 1, to maximize low end p ow er.
8. Repeat steps 6 and 7 un til the en tire band is p eak ed.
9. Record the v alue of calibration constant #238, YTM O set; Band 1 .

Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace

shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 6 through 8 again.

Change calibration constant #238 bac k to its original v alue.

Band 2 Adjustment.

10. On the syn thesizer, press:

ST ART

STOP

13.5

11. Set the oscilloscop e v ertical p osition as necessary to display the trace.
12. Mo dify calibration constant #226, YTM Gain; Band 2, to maximize high end p ow er. The

high end p ow er may p eak twice. If it do es, adjust the calibration constant to the low er

p eak (where the v alue of the calibration constant is the low er n um b er).

13. Mo dify calibration constant #239, YTM O set; Band 2, to maximize low end p ow er.

P ow er may p eak twice. Adjust to the low er p eak.

14. Adjust band 2 SRD bias calibration constants as follo ws:

a. Incremen t calibration constant #106, SRD Bias A; Band 2, to maximize low end

p ow er. Then decrease the calibration constant v alue b y 15 coun ts.

b. Incremen t calibration constant #119, SRD Bias B; Band 2, to maximize high end

p ow er. Then decrease the p ow er b y one third division ( 1/2 dB).

15. Repeat steps 12 through 14 un til the en tire band is p eak ed.
16. Record the v alue of calibration constant #239, YTM O set; Band 2 .

Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace

shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 12 through 15.

Change calibration constant #239 bac k to its original v alue.

Band 3 Adjustment.

17. On the syn thesizer, press:

ST ART

13.5

STOP

18. Set A12R25 B3S1 fully clo c kwise (see Figure 3-12).

There are tw o sets of p oten tiometers on the A12 assembly with iden tical

names. Make c ertain y ou are adjusting the correct ones.

Show 20 more labels / short text on this page
· · 7 · GHz · · · · GHz · · · · GHz · · · · 20 · · GHz · 3-26 Adjustments · Amplifier/Multiplier Adjustments
Page 286 of the Agilent Technologies 83640B user manual —  ( 83640B /L)
Page 286 · click to enlarge
Page 286
Agilent Technologies 83640B — 08360125.pdf diagram (page 286)
Fig. 95 · 08360125.pdf

Figure 3-12. A12 Assembly P otentiometer Locations

19. Set calibration constant #227, YTM Gain; Band 3, to 2048.
20. Mo dify calibration constant #240, YTM O set; Band 3, to maximize low end p ow er.

P ow er may p eak twice. Adjust to the low er p eak.

21. Adjust A12R25 B3S1 (see Figure 3-12 ) for maximum p ow er at the high end. The high

end p ow er may p eak twice. If it do es, adjust the p oten tiometer to the low er p eak (where

the adjustmen t is furthest coun terclo c kwise).

22. Set the band 3 SRD bias calibration constants as follows:

a. Incremen t calibration constant #107, SRD Bias A; Band 3, to maximize low end

p ow er. Then decrease the calibration constant v alue b y 15 coun ts.

b. Incremen t calibration constant #120, SRD Bias B; Band 3, to maximize high end

p ow er. Then decrease the p ow er b y one third division ( 1/2 dB).

23. Repeat steps 20 through 22 un til p ow er is optimized ov er the full band.
24. Record the v alue of calibration constant #240, YTM O set; Band 3 .

Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace

shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 18 through 23.

Change calibration constant #240 bac k to its original v alue.

Band 4 Adjustment (83630B/L, 83640B/L, and 83650B/L only).

25. On the syn thesizer, press:

ST ART

26.5

( 83630B /L and 83650B /L )

STOP

STOP

25.5

( 83640B /L)

26. See Figure 3-12 and set:

A12R23 B4BP1 fully clo c kwise.

A12R24 BP2 fully clo c kwise.

A12R26 B4S1 fully clo c kwise.

A12R27 S2 fully clo c kwise.

27. Set the oscilloscop e v ertical p osition as necessary to display the trace.
Show 13 more labels / short text on this page
· · 20 · · GHz · · · GHz · · · GHz · Agilent 8360 Adjustments · Amplifier/Multiplier Adjustments
Page 287 of the Agilent Technologies 83640B user manual —  ( 83650B /L)
Page 287 · click to enlarge
Page 287
28. Mo dify calibration constant #241, YTM O set; Band 4, to maximize p ow er at the v ery

low end. P ow er may p eak twice. Adjust to the low er p eak.

29. Mo dify calibration constant #228, YTM Gain; Band 4, to maximize p ow er at the rst

division on the display . P ow er may p eak twice. Adjust to the low er p eak.

If there is no noticeable increase in p ow er in the next tw o adjustmen ts, leav e

the p oten tiometers at their preset p osition (fully clo c kwise).

30. Adjust A12R26 B4S1 for maximum high-end p ow er. P ow er may p eak twice. Adjust to the

low er p eak.

31. Adjust A12R23 B4BP1 for maximum p ow er to optimize the 3rd through the 7th divisions.

Stop when the p ow er b egins to decrease. P ow er may p eak twice. Adjust to the low er

p eak.

32. Repeat step 30.
33. Adjust band 4 SRD bias calibration constants as follo ws:

a. Incremen t calibration constant #108, SRD Bias A; Band 4, to maximize low end

p ow er. Then decrease the calibration constant v alue b y 15 coun ts.

b. Incremen t calibration constant #121, SRD Bias B; Band 4, to maximize high end

p ow er. Then decrease the p ow er b y one third division ( 1/2 dB).

34. Repeat steps 28 through 33 un til the en tire band is p eak ed.
35. Record the v alue of calibration constant #241, YTM O set; Band 4 .

Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace

shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 26 through 34.

Change calibration constant #241 bac k to its original v alue.

Band 5 Adjustment (83640B/L and 83650B/L only).

36. On the syn thesizer, press:

25.5

( 83640B /L)

ST ART

26.5

( 83650B /L)

ST ART

STOP

( 83640B /L)

STOP

33.5

( 83650B /L)

37. Adjust calibration constant #229, YTM Gain; Band 5, for maximum p ow er at the high

end of band 5. The high end p ow er may p eak twice. If it do es, adjust the calibration

constant to the low er p eak (where the v alue of the calibration constant is the low er

n um b er).

38. Mo dify calibration constant #242, YTM O set; Band 5, to maximize low end p ow er.

P ow er may p eak twice. Adjust to the low er p eak.

39. Set the band 5 SRD bias calibration constants as follows:

a. Incremen t calibration constant #109, SRD Bias A; Band 5, to maximize low end

p ow er. Then decrease the calibration constant v alue b y 15 coun ts.

b. Incremen t calibration constant #122, SRD Bias B; Band 5, to maximize high end

p ow er. Then decrease the p ow er b y one third division ( 1/2 dB).

40. Repeat steps 37 through 39 un til p ow er is optimized ov er the full band.
41. Record the v alue of calibration constant #242, YTM O set; Band 5 .

Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace

shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 37 through 40.

Change calibration constant #242 bac k to its original v alue.

Show 15 more labels / short text on this page
· · GHz · · · GHz · · · 32 · GHz · · · GHz · 3-28 Adjustments · Amplifier/Multiplier Adjustments
Page 288 of the Agilent Technologies 83640B user manual — ST ART
Page 288 · click to enlarge
Page 288

Band 6 Adjustment (83640B/L and 83650B/L only).

42. On the syn thesizer, press:

( 83640B /L)

ST ART

33.5

( 83650B /L)

ST ART

STOP

( 83640B /L)

STOP

( 83650B /L)

43. Adjust calibration constant #230, YTM Gain; Band 6, for maximum p ow er at the high

end of band 6. The high end p ow er may p eak twice. If it do es, adjust the calibration

constant to the low er p eak (where the v alue of the calibration constant is the low er

n um b er).

44. Mo dify calibration constant #243, YTM O set; Band 6, to maximize low end p ow er.

P ow er may p eak twice. Adjust to the low er p eak.

45. Set the band 6 SRD bias calibration constants as follows:

a. Incremen t calibration constant #110, SRD Bias A; Band 6, to maximize low end

p ow er. Then decrease the calibration constant v alue b y 15 coun ts.

Incremen t calibration constant #123, SRD Bias B; Band 6, to maximize high end

p ow er. Then decrease the p ow er b y one third division ( 1/2 dB).

46. Repeat steps 43 through 45 un til p ow er is optimized ov er the full band.
47. Record the v alue of calibration constant #243, YTM O set; Band 6 .

Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace

shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 43 through 46.

Change calibration constant #243 bac k to its original v alue.

Band 7 Adjustment (83650B/L only).

48. On the syn thesizer, press:

ST ART

STOP

49. Adjust calibration constant #231, YTM Gain; Band 7, for maximum p ow er at the high

end of band 7. The high end p ow er may p eak twice. If it do es, adjust the calibration

constant to the low er p eak (where the v alue of the calibration constant is the low er

n um b er).

50. Mo dify calibration constant #244, YTM O set; Band 7, to maximize low end p ow er.

P ow er may p eak twice. Adjust to the low er p eak.

51. Set the band 7 SRD bias calibration constants as follows:

a. Incremen t calibration constant #111, SRD Bias A; Band 7, to maximize low end

p ow er. Then decrease the calibration constant v alue b y 15 coun ts.

b. Incremen t calibration constant #124, SRD Bias B; Band 7, to maximize high end

p ow er. Then decrease the p ow er b y one third division ( 1/2 dB).

52. Repeat steps 49 through 51 un til p ow er is optimized ov er the full band.
53. Record the v alue of calibration constant #244, YTM O set; Band 7 .

Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace

shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 49 through 52.

Change calibration constant #244 bac k to its original v alue.

Multi-band Amplifier/Multiplier Delay Compensation.

Show 27 more labels / short text on this page
· · 32 · GHz · · · GHz · · · 40 · GHz · · · 38 · GHz · · · 38 · GHz · · · · 50 · · GHz · Agilent 8360 Adjustments · Amplifier/Multiplier Adjustments
Page 289 of the Agilent Technologies 83640B user manual — SWEEP TIME
Page 289 · click to enlarge
Page 289

P ow er drop outs at the start of eac h band are typically remov ed with the YTM

Risetime calibration constants. These will b e adjusted later.

54. On the syn thesizer, press:

2.0

ST ART

STOP

SWEEP TIME

The syn thesizer in tensity marker is set to iden tify the bandcross frequency

(13.5 GHz). Reduce oscilloscop e trace in tensity to iden tify the marker. Y ou

can also iden tify bandcrosses b y temp orarily removing the cable from the

Z-AXIS BLANK/MKRS connector on the syn thesizer rear panel. Make sure

this cable is connected when making adjustmen ts.

55. On the syn thesizer, press:

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

13.5

(start of band 3)

56. Adjust calibration constant #172, YTM Bx Dly T erm A3, to maximize p ow er at the low

end of band 3 (adjust the low est amplitude p ortion of the band).

57. Adjust calibration constant #185, YTM Bx DL Y T erm B3 to maximize p ow er at the high

end of band 3.

58. Switch b et w een the fastest sweep time (0 ms) and 200 ms sweep time and c hec k that

p ow er c hanges less than 2/3 division ( 1 dB). If the c hange in p ow er is greater than 2/3

division, readjust the A term of the dela y adjustmen t #172 for low end problems, or the B

term of the dela y adjustmen t #185 for high end problems.

59. On the syn thesizer, press:

CONT

SWEEP TIME

60. Press

sev eral times to initiate sev eral sweeps. If a drop in p ow er greater than 1

SINGLE

division o ccurs when in single sweep, increase the A term of the dela y adjustmen t #172 to

put the A term at the high end of the p eak. (See \Adjustmen t Help.")

Band 4 Multi-band Amplifier/Multiplier Delay Compensation (83630B/L, 83640B/L, and 83650B/L

only).

61. On the syn thesizer, press:

2.0

ST ART

26.5

( 83630B /L and 83650B /L )

STOP

STOP

25.5

( 83640B /L)

CONT

SWEEP TIME

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

(start of band 4)

MARKER

62. Adjust calibration constant #173, YTM Bx Dly T erm A4, to maximize p ow er at the start

of band 4.

63. Adjust calibration constant #186, YTM Bx Dly T erm B4, to maximize p ow er at the high

end of band 4.

64. Switch b et w een the fastest sweep time (0 ms) and 500 ms sweep time and c hec k that

p ow er c hanges less than 2/3 division ( 1 dB) across b oth bands. If the c hange is greater

than 2/3 division, readjust the appropriate delay term.

65. On the syn thesizer, press:

SWEEP TIME

Show 52 more labels / short text on this page
· · · GHz · · · 20 · · GHz · · · 0 · · msec · · · Marker M1 · GHz · · · · 0 · msec · · · · · GHz · · · GHz · · · GHz · · · · 0 · msec · · · · Marker M3 · 20 · GHz · · · 0 · msec · · 3-30 Adjustments · Amplifier/Multiplier Adjustments
Page 290 of the Agilent Technologies 83640B user manual — 83620B /22B/24B
Page 290 · click to enlarge
Page 290
66. Press

SINGLE

sev eral times to initiate sev eral sweeps. If a drop in p ow er greater than 1

division o ccurs when in single sweep, readjust the appropriate dela y term for the lo cation

of the problem.

67. On the syn thesizer, press:

ST ART

CONT

SWEEP TIME

68. Press

SINGLE

sev eral times to initiate sev eral sweeps. If a drop in p ow er greater than 1/3

division o ccurs when in single sweep, readjust the appropriate dela y term for the lo cation

of the problem.

V erification.

69. T ry the follo wing start and stop frequencies:

T able 3-2. Start and Stop Frequencies (GHz)

83620B /22B/24B

83630B /L 83640B/L 83650B /L

83623B/L

0.01 to 20

0.01 to 26.5 0.01 to 40 0.01 to 50

4.5 to 18 4 to 23 4.5 to 32 4.5 to 33

10 to 20 10 to 20 10 to 40 10 to 50

2 to 14.5 2 to 14.5 2 to 23 2 to 23

1 F or 83622B/24B only , set from 2 GHz to 20 GHz

Compare sweeps of 0 ms and 1s and also single sweeps at fast sp eed. If a p ow er drop of

0.5 division or greater is noted, readjust the appropriate calibration constant. If problems

o ccur in other bands, adjust the appropriate YTM Bx Dly T erms for that band.

70. P erform \Ampli er/Filter Adjustmen ts" next.

Related P erformance T ests

Maxim um Lev eled P ow er

In Case Of Difficulty

1. See \Adjustmen t Help" in this pro cedure.
2. V erify the oscilloscop e ALC b oard calibration. When calibrated, the v oltage on the

oscilloscop e is scaled to 33 mV/dB.

3. Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

Show 13 more labels / short text on this page
· · 15 · GHz · · · · · 0 · msec · · Agilent 8360 Adjustments · Amplifier/Multiplier Adjustments
Page 291 of the Agilent Technologies 83640B user manual — USER CAL
Page 291 · click to enlarge
Page 291
8. Amplifier/Filter Adjustments

This pro cedure maximizes RF p ow er b y tracking the YIG tuned lter to the RF output

frequency . Initial tracking is done in single band sweeps at slow sweep sp eeds to eliminate

the e ects of bandcross and h ysteresis in the tuning coil. (Those will b e corrected with dela y

calibration constants.) Auto tracking is initiated to optimize the slow sweep tracking.

The YTF dela y comp ensation adjustmen ts maximize p ow er for fast single- and m ulti-band

sweeps. The YTM risetime adjustmen ts are done in m ulti-band fast sweeps to optimize p ow er

at the start of eac h frequency band.

A squegging clamp adjustmen t limits the p ow er and is adjusted for maximum output p ow er

without squegging. A \squegg" is a distortion in the RF output caused b y to o m uc h p ow er to

the ampli er/multiplier and c haracterized b y a p ow er drop out in a p ortion of the trace, or a

p ow er drop out ov er a broad frequency range.

If y ou ha v e replaced either the YO or the YO driver, rst initiate an auto

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Auto Track .) If auto tracking

USER CAL

passes and the instrumen t passes the \Maxim um Lev eled P ow er" p erformance

test, do not con tinue with this pro cedure. If auto tracking fails, (an error

message is display ed), con tinue with this pro cedure.

Show 6 more labels / short text on this page
Description · tracking (press · · Tracking Menu · 3-32 Adjustments · Amplifier/Filter Adjustments
Page 292 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 292 · click to enlarge
Page 292

The follo wing explanations are provided for a b etter understanding of the ampli er/ lter

adjustmen ts. Refer to these explanations as often as necessary . They will help make these

adjustmen ts easier. (The adjustmen t pro cedure follo ws.)

Offset and Gain Adjustments.

F or all o set and gain adjustmen ts, adjust thr ough the bandpass. Keep adjusting un til the

p ow er p eaks and then drops o . Then reset the adjustmen t to the p eak ed p oint.

F or all o set and gain adjustmen ts, if p ow er stays p eak ed ov er sev eral calibration constant

v alues, set the calibration constant to the middle v alue of the p eak ed range.

Squegg Clamp Calibration Constants.

Squegg clamp calibration constants are av ailable to limit the p ow er in to the

ampli er/multiplier. The band 1 adjustmen t is set for a sp eci c output p ow er at the start

of band 1. All other squegg clamp adjustmen ts are set to a default v alue to allo w maxim um

output p ow er.

When increasing an y squegg clamp calibration constant, p ow er is increased to o far when an y

of the following o ccurs:

An y p ortion of the trace is distorted or p ow er drops out in a p ortion of the trace (sphere

squegging).

P ow er drops ov er a broad frequency range.

P ow er stops increasing on an y p ortion of the trace.

Adjusting for the Center of the P assband.

The passband and tuning linearity of the ampli er/ lter v aries with frequency . The

ampli er/ lter adjustmen ts track the ampli er/ lter passband with the ampli er/multiplier

output frequency ov er the full frequency range of the syn thesizer.

Adjustment Help · Agilent 8360 Adjustments · Amplifier/Filter Adjustments

Page 293 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 293 · click to enlarge
Page 293

Single Band Delay Compensation Adjustments.

If a drop in p ow er greater than 1.5 dB o ccurs when in single sweep, increase the A dela y

term to put the A term at the high end of the p eak.

F or example, for the following calibration constant v alues:

Start of the p eak = 1800

Middle of the p eak = 1950

End of the p eak = 2100

Set the A term to 2050

See Figure 3-16 for a graphic represen tation of the frequencies a ected b y the calibration

constants.

Setting the F astest Sw eep Time.

Setting the sweep time to 0 milliseconds sets the syn thesizer to the fastest sweep time for

the frequency range b eing swept. Eac h time the frequency range is c hanged, the sweep time

m ust b e reset to 0 ms in order to maintain the fastest sweep time.

Sw eep Speed Related Adjustments.

Delay comp ensation and risetime adjustmen ts are a ected b y sweep sp eed. The adjustmen ts

are p erformed at fast sweep sp eeds whic h are the w orst case. All other adjustmen ts are

p erformed at slow sweep sp eeds; they are not a ected b y sweep sp eed.

If y ou ha v e a sweep sp eed problem, it is probably a ected b y dela y comp ensation or risetime

adjustmen ts.

YTF Bandcross Delay T erms.

The YTF Bx Dly terms are either o set or gain terms. A1 denotes the o set term for band 1.

B1 denotes the gain term for band 1. The o set (A) should b e adjusted to maximize p ow er at

the b eginning of the band. The gain (B) is adjusted to maximize p ow er tow ard the end of the

band.

See Figure 3-16 and Figure 3-17 for a graphic represen tation of the frequencies a ected b y the

calibration constants.

3-34 Adjustments · Amplifier/Filter Adjustments

Page 294 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 294 · click to enlarge
Page 294
Agilent Technologies 83640B — 08360125.pdf diagram (page 294)
Fig. 96 · 08360125.pdf

Figure 3-16. Single-Band Delay and Risetime Compensation

Agilent 8360 Adjustments · Amplifier/Filter Adjustments

Page 295 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 295 · click to enlarge
Page 295
Agilent Technologies 83640B — 08360125.pdf diagram (page 295)
Fig. 97 · 08360125.pdf

Figure 3-17. Multi-Band Delay and Risetime Compensation

3-36 Adjustments · Amplifier/Filter Adjustments

Page 296 of the Agilent Technologies 83640B user manual — SERVICE
Page 296 · click to enlarge
Page 296
1. Connect the equipmen t as sho wn in Figure 3-18 using an analog oscilloscop e with A

v ersus B sweep capabilit y . Do not connect the p ow er meter y et. (See Figure 3-19 for

lo cation detail of XA10J2 pin 55 on the motherb oard.) P ow er on all the instrumen ts and

let them w arm up for at least one hour.

2. Press

PRESET

on the syn thesizer and, for the 83640B /L and 83650B /L press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Disable Doubler (asterisk on)

SERVICE

3. On the p ow er meter:

Zero and calibrate the p ow er meter/sensor.

Set the p ow er meter to dBm mo de.

Agilent Technologies 83640B — SERVICE diagram (page 296)
Fig. 98

Figure 3-18. Amplifier/Filter Adjustments Setup

Analog Oscilloscop e HP/Agilent 1740A

Oscilloscop e Prob es HP/Agilent 10431A

P ow er Meter HP/Agilent 436A/7A/8A

P ow er Sensor HP/Agilent 8485A (83620B/22B/24B,

P ow er Sensor HP/Agilent 8487A (83640B/L and 83650B /L)

A tten uator 10 dB HP/Agilent 8493C Option 010 (83620B /22B/24B,

A tten uator 10 dB HP/Agilent 8490D Option 010 (83640B /L and 83650B/L)

Show 9 more labels / short text on this page
Procedure · · · Tools Menu · Equipment · 83623B/L, and 83630B/L) · 83623B/L, and 83630B/L) · Agilent 8360 Adjustments · Amplifier/Filter Adjustments
Page 297 of the Agilent Technologies 83640B user manual — ENTER
Page 297 · click to enlarge
Page 297
Agilent Technologies 83640B — 08360125.pdf diagram (page 297)
Fig. 99 · 08360125.pdf

Figure 3-19. XA10J2 Location

4. Set the default v alues given in T able 3-4 in to the corresp onding ampli er/ lter

calibration constants. On the syn thesizer, set:

SERVICE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Select Cal En ter the n um b er of the rst calibration constant from T able 3-4 and

terminate the en try with

. When using the front panel em ulator softw are,

ENTER

pressing the arrow k eys will allo w y ou to v ary the calibration constants. The calibration

constant sho wn at the top of the \display" window is the activ e constant.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Modify Cal En ter the default v alue from T able 3-4 and terminate the en try with

.

ENTER

Using the up/do wn arrow k eys, select eac h of the rest of the calibration constants in

T able 3-4 and set their default v alues b y en tering the v alue on the n umeric k eypad.

T erminate y our en try with the

ENTER

k ey .

Show 7 more labels / short text on this page
· · Adjust Menu · Calib Menu · · 3-38 Adjustments · Amplifier/Filter Adjustments
Page 298 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 298 · click to enlarge
Page 298

T able 3-4. Amplifier/Filter Calibration Constants and Default V alues

Numb er Description Default

131 Squegg Clamp 1A 255 Maximize p ow er across the

band without squegging.

132 Squegg Clamp 2A 255

Adjusted at slow sw eep sp eeds

133 Squegg Clamp 3A 255

and single sw eep mo de where the

134 Squegg Clamp 4A 255

SYTM sphere is most likely to

squegg.

135 Squegg Clamp 5A 255

136 Squegg Clamp 6A 255

137 Squegg Clamp 7A 255

205 YTM Rise; Band 1 15 Minimize p ow er drop outs at the

start of band. Adjustmen t is only

206 YTM Rise; Band 2 A1 20

e ectiv e at fast sw eep sp eeds.

207 YTM Rise; Band 2 B1 2

208 YTM Rise; Band 2 A2 15

209 YTM Rise; Band 3 A1 70

210 YTM Rise; Band 3 B1 7

211 YTM Rise; Band 3 A2 15

212 YTM Rise; Band 3 B2 15

213 YTM Rise; Band 3 A3 15

214 YTM Rise; Band 4 15

215 YTM Rise; Band 5 15

216 YTM Rise; Band 6 10

217 YTM Rise; Band 7 10

Show 5 more labels / short text on this page
Adjustmen t · V alue · Description · Agilent 8360 Adjustments · Amplifier/Filter Adjustments
Page 299 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 299 · click to enlarge
Page 299

T able 3-4.

Amplifier/Filter Calibration Constants and Default V alues (continued)

Numb er Description Default

249 Autotrack D A C Setting 2048 Sets p ow er level during

Autotrack or P ow er P eaking

449 D Amp Clamp 1 255 Protect the bridge detector

from excessive p ow er when

450 D Amp Clamp 2 255

doubler amp mo de is selected.

451 D Amp Clamp 3 255

452 D Amp Clamp 4 255

453 D Amp Clamp 5 255

454 D Amp Clamp 6 255

455 D Amp Clamp 7 255

561 YTF Delay T erm A Hrm 1 1000 Maximize p ow er over rst 20%

of band. A ects fast sw eeps

562 YTF Delay T erm A Hrm 2 1000

only . Single sw eep mo de is also

563 YTF Delay T erm A Hrm 3 1000

critical.

564 YTF Delay T erm A Hrm 4 1000

565 YTF Delay T erm A Hrm 5 1000

566 YTF Delay T erm A Hrm 6 1000

567 YTF Delay T erm A Hrm 7 1000

571 YTF Delay T erm B Hrm 1 300 Maximize p ow er over higher 30%

572 YTF Delay T erm B Hrm 2 300

of band. A ects fast sw eeps only .

573 YTF Delay T erm B Hrm 3 300

574 YTF Delay T erm B Hrm 4 300

575 YTF Delay T erm B Hrm 5 300

576 YTF Delay T erm B Hrm 6 300

577 YTF Delay T erm B Hrm 7 300

591 YTF Bx Dly T erm A1 1000 Delay comp ensation for m ulti-

592 YTF Bx Dly T erm A2 1000

band sw eeps only . Adjust for

maxim um p ow er at low er 50% of

593 YTF Bx Dly T erm A3 1000

band. A ects fast sw eep sp eeds

594 YTF Bx Dly T erm A4 1000

only .

595 YTF Bx Dly T erm A5 1000

596 YTF Bx Dly T erm A6 1000

597 YTF Bx Dly T erm A7 1000

Show 5 more labels / short text on this page
Adjustment · V alue · Description · 3-40 Adjustments · Amplifier/Filter Adjustments
Page 300 of the Agilent Technologies 83640B user manual — ALC
Page 300 · click to enlarge
Page 300

T able 3-4.

Amplifier/Filter Calibration Constants and Default V alues (continued)

Numb er Description Default

601 YTF Bx Dly T erm B1 300 Delay comp ensation for m ulti-

band sw eeps only . Adjust for

602 YTF Bx Dly T erm B2 300

maxim um p ow er at higher 30% of

603 YTF Bx Dly T erm B3 300

band. A ects fast sw eep sp eeds

only .

604 YTF Bx Dly T erm B4 300

605 YTF Bx Dly T erm B5 300

606 YTF Bx Dly T erm B6 300

607 YTF Bx Dly T erm B7 300

621 *YTF Gain Band 1 2048 Maximize p ow er over higher 10%

of band. Adjust at slow sw eep

622 *YTF Gain Band 2 2048

sp eeds.

623 *YTF Gain Band 3 2048

624 *YTF Gain Band 4 2048

625 *YTF Gain Band 5 2048

626 *YTF Gain Band 6 2048

627 *YTF Gain Band 7 2048

631 *YTF O set Band 1 2048 Maximize p ow er over low er 10%

of band. Adjust at slow sw eep

632 *YTF O set Band 2 2048

sp eeds.

633 *YTF O set Band 3 2048

634 *YTF O set Band 4 2048

635 *YTF O set Band 5 2048

636 *YTF O set Band 6 2048

637 *YTF O set Band 7 2048

Amplifier/Filter Offset and Gain Adjustment

Band 1 Adjustment.

5. On the syn thesizer, set:

ST ART

2.0

STOP

SWEEP TIME

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Leveling Mode ALCoff (asterisk on)

ALC

POWER LEVEL

6. On the oscilloscop e, set:

Channel A:

Volts/Division5 mV/Division
Volts/Division10 mV/Division (83623B/L and 83624B only)
OffsetAs required
Input CouplingDC
Input Impedance1 M
Volts/Division1 V/Division
OffsetAs required
Show 27 more labels / short text on this page
Adjustmen t · V alue · Description · · · GHz · · · · 7 · GHz · · · · 200 · msec · · · · · · 25 · dBm · · Channel B · Agilent 8360 Adjustments · Amplifier/Filter Adjustments
Page 301 of the Agilent Technologies 83640B user manual — SWEEP
Page 301 · click to enlarge
Page 301
Input CouplingDC
Input Impedance1 M
Sweep ModeA versus B

Adjust Channel A o set, v olts/division, and horizontal p osition con trols for a trace that

lls the full horizontal display . Adjust the v ertical p osition with the A v ertical p osition

con trol.

In the following adjustmen t p ow er drop outs may o ccur due to squegging.

Ignore the squegging (the squegg clamp will b e adjusted later) and p eak the

adjustmen t for maximum p ow er.

7. Decrease the syn thesizer p ow er lev el un til the lev el of the en tire trace on the oscilloscop e

b egins to drop.

8. Adjust A12R126 YTFG to maximize high end p ow er (last tw o horizontal display

divisions).

9. Mo dify calibration constant #631, YTF O set; Band 1, to maximize low end p ow er.
10. Repeat steps 8 and 9 un til the en tire band is p eak ed.
11. Record the v alue of calibration constant #631, YTF O set; Band 1 .

Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace

shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 7 through 10 un til it

is. Then c hange calibration constant #631 bac k to the recorded v alue.

12. On the syn thesizer, set:

POWER LEVEL

13. Adjust calibration constant #131, Squegg Clamp 1A, to maximize RF output p ow er

across the band. The clamp is adjusted correctly when the follo wing is true:

a. An increase in the calibration constant v alue decreases the output p ow er.

b. A small decrease in the calibration constant v alue results in no c hange in output

p ow er.

14. If y ou ha v e an 83623B or an 83624B , con tinue with \Band 2 Adjustmen t." If y ou ha v e

an 83623L, con tinue with the next step. Otherwise, adjust calibration constant #131,

Squegg Clamp 1A as describ ed here, to protect the bridge detector from excessiv e p ow er:

a. Set the oscilloscop e v ertical p osition as necessary to display the trace.

b. On the syn thesizer, set:

SWEEP

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Manual Sweep (asterisk on)

c. Using the rotary knob, nd the p eak of the sweep as display ed b y the oscilloscop e.

d. If the p ow er lev el of the p eak p oint is greater than 21.5 dBm for 83620B /22B and

83630B /L or 23.5 dBm for 83640B /L and 83650B /L, as display ed on the p ow er meter,

decrease Squegg Clamp 1A un til the p ow er lev el is less than these v alues.

If the p ow er lev el of the p eak p oint is already less than these v alues, return the

syn thesizer to con tinuous sweep and decrease Squegg Clamp 1A just un til the p eak

p oint b egins to drop (this indicates that p ow er to the bridge detector is clamp ed).

15. If y ou ha v e an 83623L, adjust calibration constant #131, Squegg Clamp 1A as describ ed

here. Otherwise, con tinue with the next step.

a. Set the syn thesizer's sweep time to 500 ms.

b. Adjust the calibration constant #131, Squegg Clamp 1A for maximum p ow er without

squegging.

c. Chec k for squegging b y setting the syn thesizer's sweep time to 0 ms. If the p ow er

increase is greater 2 dB in an y part of the band, then squegging is o ccurring.

Show 8 more labels / short text on this page
· · 0 · · dBm · · 3-42 Adjustments · Amplifier/Filter Adjustments
Page 302 of the Agilent Technologies 83640B user manual — MENU
Page 302 · click to enlarge
Page 302

d. If squegging o ccurs, readjust calibration constant #131 so that the p ow er di erence is

2 dB or less from a 500 ms (slo w) sweep to a 0 ms (fast) sweep.

16. If y ou ha v e an y instrumen t exc ept an 83640B /L and 83650B /L, con tinue with \Band 2

Adjustmen t." Otherwise, adjust calibration constant #449, DAmp Clamp 1 as describ ed

here, to protect the bridge detector from excessiv e p ow er while in doubler amp mo de:

a. On the syn thesizer, set:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Disable Doubler (asterisk o )

SERVICE

PO WER

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Doubler Amp Mode On (asterisk on)

b. Set the oscilloscop e v ertical p osition as necessary to display the trace.

c. On the syn thesizer, set:

SWEEP

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Manual Sweep (asterisk on)

d. Using the syn thesizer rotary knob, nd the p eak of the sweep as display ed b y the

oscilloscop e.

e. If the p ow er lev el of the p eak p oint is greater than 23.5 dBm, as display ed on the

p ow er meter, decrease DAmp Clamp 1 un til the p ow er lev el is less than this lev el.

If the p ow er lev el of the p eak p oint is already less than this lev el, set the syn thesizer

bac k to con tinuous sweep and decrease DAmp Clamp 1 just un til the p eak p oint

b egins to drop (this indicates that p ow er to the bridge detector is clamp ed).

17. Disable the doubler, turn doubler amp mo de o , and return the syn thesizer to con tinuous

sweep if y ou are still in man ual sweep.

Band 2 Adjustment.

18. On the syn thesizer, set:

ST ART

STOP

13.5

POWER LEVEL

19. Set the oscilloscop e v ertical p osition as necessary to display the trace.
20. Decrease the syn thesizer p ow er lev el un til the lev el of the en tire trace on the oscilloscop e

b egins to drop.

21. Mo dify calibration constant #622, YTF Gain; Band 2, to maximize high end p ow er.
22. Mo dify calibration constant #632, YTF O set; Band 2, to maximize low end p ow er.
23. Rep eat steps 20 and 21 un til the en tire band is p eak ed.
24. Record the v alue of calibration constant #632, YTF O set; Band 2 .

Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace

shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 19 through 22 un til

it is. Then c hange calibration constant #632 bac k to the recorded v alue.

25. On the syn thesizer, set:

POWER LEVEL

26. Adjust calibration constant #132, Squegg Clamp 2A, to maximize RF output p ow er

across the band.

27. If y ou ha v e an 83623B or an 83624B , con tinue with \Band 3 Adjustmen t." If y ou ha v e

an 83623L, con tinue with the next step. Otherwise, adjust calibration constant #132,

Squegg Clamp 2A as describ ed here, to protect the bridge detector from excessiv e p ow er:

a. Set the oscilloscop e v ertical p osition as necessary to display the trace.

b. On the syn thesizer, set:

SWEEP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Manual Sweep (asterisk on)

MENU

c. Using the rotary knob, nd the p eak of the sweep as display ed b y the oscilloscop e.

Show 28 more labels / short text on this page
· · Tools Menu · · Dblr Amp Menu · · · · 7 · GHz · · · · GHz · · · · 25 · dBm · · · · 0 · dBm · · · Agilent 8360 Adjustments · Amplifier/Filter Adjustments
Page 303 of the Agilent Technologies 83640B user manual — STOP
Page 303 · click to enlarge
Page 303

d. If the p ow er lev el of the p eak p oint is greater than 20 dBm for 83620B /22B and

83630B /L or 22.7 dBm for 83640B /L and 83650B /L, as display ed on the p ow er meter,

decrease Squegg Clamp 2A un til the p ow er lev el is less than these v alues.

If the p ow er lev el of the p eak p oint is already less than these v alues, return the

syn thesizer to con tinuous sweep and decrease Squegg Clamp 2A just un til the p eak

p oint b egins to drop (this indicates that p ow er to the bridge detector is clamp ed).

28. If y ou ha v e an 83623L, adjust calibration constant #132, Squegg Clamp 2A as describ ed

here. Otherwise, con tinue with the next step.

a. Set the syn thesizer's sweep time to 500 ms.

b. Adjust the calibration constant #132, Squegg Clamp 2A for maximum p ow er without

squegging.

c. Chec k for squegging b y setting the syn thesizer's sweep time to 0 ms. If the p ow er

increase is greater 2 dB in an y part of the band, then squegging is o ccurring.

d. If squegging o ccurs, readjust calibration constant #132 so that the p ow er di erence is

2 dB or less from a 500 ms (slow) sweep to a 0 ms (fast) sweep.

29. If y ou ha v e an y instrumen t exc ept an 83640B /L and 83650B /L, con tinue with \Band 3

Adjustmen t." Otherwise, adjust calibration constant #450, DAmp Clamp 2 as describ ed

here, to protect the bridge detector from excessiv e p ow er while in doubler amp mo de:

a. On the syn thesizer, set:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Disable Doubler (asterisk o )

SERVICE

PO WER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Doubler Amp Mode On (asterisk on)

MENU

b. Set the oscilloscop e v ertical p osition as necessary to display the trace.

c. On the syn thesizer, set:

SWEEP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Manual Sweep (asterisk on)

MENU

d. Using the syn thesizer rotary knob, nd the p eak of the sweep as display ed b y the

oscilloscop e.

e. If the p ow er lev el of the p eak p oint is greater than 22.7 dBm, as display ed on the

p ow er meter, decrease DAmp Clamp 2 un til the p ow er lev el is less than these v alues.

If the p ow er lev el of the p eak p oint is already less than these v alues, set the

syn thesizer bac k to con tinuous sweep and decrease DAmp Clamp 2 just un til the p eak

p oint b egins to drop (this indicates that p ow er to the bridge detector is clamp ed).

30. Disable the doubler, turn doubler amp mo de o , and return the syn thesizer to con tinuous

sweep if y ou are still in man ual sweep.

Band 3 Adjustment.

31. On the syn thesizer, set:

ST ART

13.5

STOP

POWER LEVEL

32. Decrease the syn thesizer p ow er lev el un til the lev el of the en tire trace on the oscilloscop e

b egins to drop.

33. Set A12R108 B3S1 fully clo c kwise (see Figure 3-20).

There are tw o sets of p oten tiometers on the A12 assembly with iden tical

names. Make c ertain y ou are adjusting the correct ones.

Show 22 more labels / short text on this page
· · Tools Menu · · Dblr Amp Menu · · · · GHz · · · · 20 · GHz · · · · 25 · dBm · · 3-44 Adjustments · Amplifier/Filter Adjustments
Page 304 of the Agilent Technologies 83640B user manual — MENU
Page 304 · click to enlarge
Page 304
Agilent Technologies 83640B — 08360125.pdf diagram (page 304)
Fig. 100 · 08360125.pdf

Figure 3-20. A12 Assembly P otentiometer Locations

34. Set calibration constant #623, YTF Gain; Band 3, to 2084.
35. Mo dify calibration constant #633, YTF O set; Band 3, to maximize low end p ow er.
36. Adjust A12R108 B3S1 (see Figure 3-20) for maximum p ow er ov er the rest of the sweep

(3 to 10 divisions).

37. Rep eat steps 33 and 34 un til p ow er is optimized ov er the full band.
38. Record the v alue of calibration constant #633, YTF O set; Band 3 .

Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace

shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 31 through 35 un til

it is. Then c hange calibration constant #633 bac k to the recorded v alue.

39. On the syn thesizer, set:

POWER LEVEL

40. Adjust calibration constant #133, Squegg Clamp 3A, to maximize RF output p ow er

across the band.

41. If y ou ha v e an 83623B /L or an 83624B , con tinue with \Band 4 Adjustmen t." Otherwise,

adjust calibration constant #133, Squegg Clamp 3A as describ ed here, to protect the

bridge detector from excessiv e p ow er:

a. Set the oscilloscop e v ertical p osition as necessary to display the trace.

b. On the syn thesizer, set:

SWEEP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Manual Sweep (asterisk on)

MENU

c. Using the rotary knob, nd the p eak of the sweep as display ed b y the oscilloscop e.

d. If the p ow er lev el of the p eak p oint is greater than 21.5 dBm for 83620B /22B and

83630B /L or 22.3 dBm for 83640B /L and 83650B /L, as display ed on the p ow er meter,

decrease Squegg Clamp 3A un til the p ow er lev el is less than these v alues.

If the p ow er lev el of the p eak p oint is already less than these v alues, return the

syn thesizer to con tinuous sweep and decrease Squegg Clamp 3A just un til the p eak

p oint b egins to drop

(this indicates that p ow er to the bridge detector is clamp ed).

Show 8 more labels / short text on this page
· · 0 · dBm · · · Agilent 8360 Adjustments · Amplifier/Filter Adjustments
Page 305 of the Agilent Technologies 83640B user manual — A12R109 B4S1
Page 305 · click to enlarge
Page 305
42. If y ou ha v e an y instrumen t exc ept an 83640B /L and 83650B /L, con tinue with \Band 4

Adjustmen t." Otherwise, adjust calibration constant #451, DAmp Clamp 3 as describ ed

here, to protect the bridge detector from excessiv e p ow er while in doubler amp mo de:

a. On the syn thesizer, set:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SERVICE

Disable Doubler (asterisk o )

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PO WER

MENU

Doubler Amp Mode On (asterisk on)

b. Set the oscilloscop e v ertical p osition as necessary to display the trace.

c. On the syn thesizer, set:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SWEEP

MENU

Manual Sweep (asterisk on)

d. Using the syn thesizer rotary knob, nd the p eak of the sweep as display ed b y the

oscilloscop e.

e. If the p ow er lev el of the p eak p oint is greater than 22.3 dBm, as display ed on the

p ow er meter, decrease DAmp Clamp 3 un til the p ow er lev el is less than these v alues.

If the p ow er lev el of the p eak p oint is already less than these v alues, set the

syn thesizer bac k to con tinuous sweep and decrease DAmp Clamp 3 just un til the p eak

p oint b egins to drop (this indicates that p ow er to the bridge detector is clamp ed).

43. Disable the doubler, turn doubler amp mo de o , and return the syn thesizer to con tinuous

sweep if y ou are still in man ual sweep.

Band 4 Adjustment (83630B/L, 83640B/L, and 83650B/L only).

44. On the syn thesizer, set:

ST ART

26.5

( 83630B /L and 83650B /L )

STOP

25.5

( 83640B /L)

STOP

POWER LEVEL

45. Decrease the syn thesizer p ow er lev el un til the lev el of the en tire trace on the oscilloscop e

b egins to drop.

46. Set the follo wing p oten tiometers, lo cated on the left side of the b oard, fully clo c kwise:

A12R109 B4S1

A12R110 S2

A12R105 B4BP1

A12R106 BP2

47. Set the oscilloscop e v ertical p osition as necessary to display the trace.
48. Mo dify calibration constant #634, YTF O set; Band 4, to maximize low end p ow er.
49. Mo dify calibration constant #624, YTF Gain; Band 4, to maximize p ow er at the rst

division on the display .

If there is no noticeable increase in p ow er in the next tw o adjustmen ts, leav e

the p oten tiometers at their preset p osition (fully clo c kwise).

50. Adjust A12R109 B4S1 for maximum high-end p ow er.
51. Adjust A12R105 B4BP1 for maximum p ow er to optimize the 3rd through the 7th

divisions. Stop when the p ow er b egins to decrease.

52. Repeat step 47.
53. Repeat steps 41 through 48 un til the en tire band is p eak ed.
54. Record the v alue of calibration constant #634, YTM O set; Band 4 .
Show 24 more labels / short text on this page
· · Tools Menu · · Dblr Amp Menu · · · · 20 · · GHz · · · GHz · · · GHz · · · 25 · dBm · · 3-46 Adjustments · Amplifier/Filter Adjustments
Page 306 of the Agilent Technologies 83640B user manual —  ( 83650B /L)
Page 306 · click to enlarge
Page 306

Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace

shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 46 through 51 un til

it is. Then c hange calibration constant #634 bac k to the recorded v alue.

55. On the syn thesizer, set:

POWER LEVEL

56. Adjust calibration constant #134, Squegg Clamp 4A , to maximize RF output p ow er

across the band.

57. F or the 83630B /L only , adjust calibration constant #134, Squegg Clamp 4A as describ ed

here, to protect the bridge detector from excessiv e p ow er:

a. Set the oscilloscop e v ertical p osition as necessary to display the trace.

b. On the syn thesizer, set:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SWEEP

MENU

Manual Sweep (asterisk on)

c. Using the rotary knob, nd the p eak of the sweep as display ed b y the oscilloscop e.

d. If the p ow er lev el of the p eak p oint is greater than 18.9 dBm, as display ed on the

p ow er meter, decrease Squegg Clamp 4A un til the p ow er lev el is less than these

v alues.

If the p ow er lev el of the p eak p oint is already less than these v alues, return the

syn thesizer to con tinuous sweep and decrease Squegg Clamp 4A just un til the p eak

p oint b egins to drop

(this indicates that p ow er to the bridge detector is clamp ed).

58. F or the 83640B /L and 83650B /L only , adjust calibration constant #452,

DAmp Clamp 4 as describ ed here, to protect the bridge detector from excessiv e p ow er

while in doubler amp mo de:

a. On the syn thesizer, set:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Disable Doubler (asterisk o )

SERVICE

PO WER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Doubler Amp Mode On (asterisk on)

MENU

b. Set the oscilloscop e v ertical p osition as necessary to display the trace.

c. On the syn thesizer, set:

SWEEP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Manual Sweep (asterisk on)

MENU

d. Using the syn thesizer rotary knob, nd the p eak of the sweep as display ed b y the

oscilloscop e.

e. If the p ow er lev el of the p eak p oint is greater than 23.5 dBm, as display ed on the

p ow er meter, decrease DAmp Clamp 4 un til the p ow er lev el is less than these v alues.

If the p ow er lev el of the p eak p oint is already less than these v alues, set the

syn thesizer bac k to con tinuous sweep and decrease DAmp Clamp 4 just un til the p eak

p oint b egins to drop (this indicates that p ow er to the bridge detector is clamp ed).

59. Disable the doubler, turn doubler amp mo de o , and return the syn thesizer to con tinuous

sweep if y ou are still in man ual sweep.

Band 5 Adjustment (83640B/L and 83650B/L only).

60. On the syn thesizer set:

ST ART

25.5

( 83640B /L)

ST ART

26.5

( 83650B /L)

STOP

( 83640B /L)

33.5

( 83650B /L)

STOP

POWER LEVEL

61. Set the oscilloscop e v ertical p osition as necessary to display the trace.
62. Decrease the syn thesizer p ow er lev el un til the lev el of the en tire trace on the oscilloscop e

b egins to drop.

Show 32 more labels / short text on this page
· · 0 · dBm · · · · · Tools Menu · · Dblr Amp Menu · · · · GHz · · · GHz · · · 32 · GHz · · · GHz · · · 25 · · dBm · Agilent 8360 Adjustments · Amplifier/Filter Adjustments
Page 307 of the Agilent Technologies 83640B user manual — ST ART
Page 307 · click to enlarge
Page 307
63. Mo dify calibration constant #625, YTF Gain; Band 5, to maximize high end p ow er.
64. Mo dify calibration constant #635, YTF O set; Band 5, to maximize low end p ow er.
65. Repeat steps 61 and 62 un til the en tire band is p eak ed.
66. Note the v alue of calibration constant #635, YTF O set; Band 5 .

Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace

shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 61 through 63 un til

it is. Then c hange calibration constant #635 bac k to the recorded v alue.

67. On the syn thesizer, set:

POWER LEVEL

68. Adjust calibration constant #135, Squegg Clamp 5A, to maximize RF output p ow er

across the band.

Band 6 Adjustment (83640B/L and 83650B/L only).

69. On the syn thesizer set:

( 83640B /L)

ST ART

ST ART

33.5

( 83650B /L)

STOP

( 83640B /L)

STOP

( 83650B /L)

POWER LEVEL

70. Set the oscilloscop e v ertical p osition as necessary to display the trace.
71. Decrease the syn thesizer p ow er lev el un til the lev el of the en tire trace on the oscilloscop e

b egins to drop.

72. Mo dify calibration constant #626, YTF Gain; Band 6, to maximize high end p ow er.
73. Mo dify calibration constant #636, YTF O set; Band 6, to maximize low end p ow er.
74. Repeat steps 70 and 71 un til the en tire band is p eak ed.
75. Note the v alue of calibration constant #636, YTF O set; Band 6 .

Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace

shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 70 through 72 un til

it is. Then c hange calibration constant #636 bac k to the recorded v alue.

76. On the syn thesizer, set:

POWER LEVEL

77. Adjust calibration constant #136, Squegg Clamp 3A, to maximize RF output p ow er

across the band.

Band 7 Adjustment (83650B/L only).

78. On the syn thesizer set:

ST ART

STOP

POWER LEVEL

79. Set the oscilloscop e v ertical p osition as necessary to display the trace.
80. Decrease the syn thesizer p ow er lev el un til the lev el of the en tire trace on the oscilloscop e

b egins to drop.

81. Mo dify calibration constant #627, YTF Gain; Band 7, to maximize high end p ow er.
82. Mo dify calibration constant #637, YTF O set; Band 7, to maximize low end p ow er.
83. Repeat steps 79 and 80 un til the en tire band is p eak ed.
84. Note the v alue of calibration constant #637, YTF O set; Band 7 .

Step do wn the calibration constant in 100 coun t steps and v erify that the en tire trace

shifts do wn ev enly . If the p ow er c hange is not uniform, rep eat steps 79 through 81 un til

it is. Then c hange calibration constant #637 bac k to the recorded v alue.

Show 47 more labels / short text on this page
· · 0 · dBm · · · · 32 · GHz · · · GHz · · · 40 · GHz · · · 38 · GHz · · · 25 · dBm · · · · 0 · dBm · · · · 38 · GHz · · · · 50 · GHz · · · · 25 · · dBm · 3-48 Adjustments · Amplifier/Filter Adjustments
Page 308 of the Agilent Technologies 83640B user manual — POWER LEVEL
Page 308 · click to enlarge
Page 308
85. On the syn thesizer, set:

POWER LEVEL

86. Adjust calibration constant #137, Squegg Clamp 7A, to maximize RF output p ow er

across the band.

Auto Tracking V erification.

87. Record the v alue of the follo wing calibration constants:

#621 YTF Gain; Band 1

#622 YTF Gain; Band 2

#623 YTF Gain; Band 3

#624 YTF Gain; Band 4

#625 YTF Gain; Band 5

#626 YTF Gain; Band 6

#627 YTF Gain; Band 7

#631 YTF O set; Band 1

#632 YTF O set; Band 2

#633 YTF O set; Band 3

#634 YTF O set; Band 4

#635 YTF O set; Band 5

#636 YTF O set; Band 6

#637 YTF O set; Band 7

#225 YTM Gain; Band 1

#226 YTM Gain; Band 2

#227 YTM Gain; Band 3

#228 YTM Gain; Band 4

#229 YTM Gain; Band 5

#230 YTM Gain; Band 6

#231 YTM Gain; Band 7

#238 YTM O set; Band 1

#239 YTM O set; Band 2

#240 YTM O set; Band 3

#241 YTM O set; Band 4

#242 YTM O set; Band 5

#243 YTM O set; Band 6

#244 YTM O set; Band 7

88. Initiate auto tracking on the syn thesizer as follo ws. T erminate the RF OUTPUT with a

go o d 50 imp edance match suc h as a 10 dB attenuator or a p ow er sensor (not required

for syn thesizers with a step attenuator). Press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USER CAL

W ait for the syn thesizer to complete auto tracking. If auto tracking fails (an error

message is display ed) rep eat the gain and o set adjustmen ts for the failed frequency .

89. On the syn thesizer, set:

ST ART

2.0

STOP

CONT

SWEEP TIME

1.5

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ALC

Leveling Point Intrnl (asterisk on)

POWER LEVEL

Show 32 more labels / short text on this page
· · 0 · · dBm · · · Tracking Menu · Auto Track · · · GHz · · · · 20 · GHz · · · · · sec · · · · · · 25 · dBm · · Agilent 8360 Adjustments · Amplifier/Filter Adjustments
Page 309 of the Agilent Technologies 83640B user manual — ST ART
Page 309 · click to enlarge
Page 309

F or the 83640B /L and 83650B /L only , also press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Disable Doubler (asterisk o )

SERVICE

Connect the p ow er meter to the syn thesizer RF OUTPUT through the p ow er sensor and

10 dB attenuator (see Figure 3-18). Adjust the oscilloscop e horizontal con trol to place

the low p ow er p oint on a v ertical graticule and note the p osition.

Select man ual sweep. Press SWEEP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Manual Sweep (asterisk on). Use the arrow

MENU

k eys to adjust the CW frequency to the minimum p ow er p oint on the oscilloscop e.

On the p ow er meter, c hec k that the p ow er at this p oint is not less than the sp eci ed

maximum lev eled p ow er (note that p ow er is attenuated b y 10 dB ). If it is, note the lev el

and frequency , and c hec k the follo wing:

a. Compare the man ually adjusted YTF and YTM gain and o set calibration constants

(step 85) to the auto track ed calibration constants. A large di erence in v alues

(several h undred coun ts) indicates a p ossible misadjustment. The auto tracking v alues

are correct.

b. If the band 3 v alues are within 500 coun ts of the limits of their ranges, rep eat the

B3S1, B4S1, and B4BP1 adjustmen ts with the doubler disabled.

c. If the band 4 v alues are within 500 coun ts of the limits of the range, rep eat the B4S1

and B4BP1 adjustmen ts with the doubler disabled (unless y ou ha v e just readjusted

these adjustmen ts in step b).

d. If the problem con tinues, it is most likely hardw are-related. Refer to the A gilent

T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series Swept CW

Gener ator T r oublesho oting Guide .

90. F or the 83630B /L, 83640B /L, and 83650B /L, rep eat step 87 for eac h frequency range

listed in T able 3-5 and a sweep time of 1.5 seconds.

T able 3-5. Frequency Ranges

83630B /L 83640B/L 83650B/L

20 to 26.5 GHz 20 to 25.5 GHz 20 to 26.5 GHz

25.5 to 40 GHz 26.5 to 38 GHz

38 to 50 GHz

91. On the syn thesizer, set:

2.0

ST ART

STOP

CONT

SWEEP TIME

1.5

92. Set the syn thesizer p ow er lev el to the sp eci ed maxim um lev eled p ow er. Chec k for ALC

oscillations or lev el squegging.

a. The SRD bias A or B calibration constants for the a ected band might eliminate

oscillation. The problem is with the A term if the oscillations are near the b eginning

of the band. It is with the B term if the oscillations are near the end of the band.

SRD bias has a range. It can b e ov erbiased whic h causes lev el squegging. If so,

decrease the v alue of the SRD bias calibration constants and rep eat the adjustmen t.

b. If the SRD bias calibration constants ha v e no e ect, c hec k the ALC Mo d Gain

calibration constant for the appropriate band.

93. F or the 83630B /L, 83640B /L, and 83650B /L, rep eat step 90 for eac h frequency range

listed in T able 3-5 and a sweep time of 1.5 seconds.

Show 20 more labels / short text on this page
· · Tools Menu · · · · · GHz · · · 20 · GHz · · · · · sec · · 3-50 Adjustments · Amplifier/Filter Adjustments
Page 310 of the Agilent Technologies 83640B user manual — SERVICE
Page 310 · click to enlarge
Page 310
94. Set the syn thesizer p ow er lev el to 25 dBm and then set v arious frequency ranges suc h as

5 GHz to 50 GHz or 10 GHz to 45 GHz.

Set the sweep times to greater than 200 ms for single-band sweeps and less than 1.5

seconds for m ulti-band sweeps since dela y comp ensation has not y et b een adjusted.

F or eac h frequency range set, compare the signal lev els of single sweep to con tinuous

sweep. Band frequency ranges are as follo ws:

83640B/L

83650B /L

0 0.01 to < 2.0 0.01 to < 2.0 0.01 to < 2.0

1 2.0 to 7.0 2.0 to 7.0 2.0 to 7.0

2 7.0 to < 13.5 7.0 to < 13.5 7.0 to < 13.5

3 13.5 to 20.0 13.5 to 20.0 13.5 to 20.0

4 > 20.0 to 26.5 > 20.0 to 25.5 > 20.0 to 26.5

5 25.5 to < 32 26.5 to < 33.5

6 32 to < 40 33.5 to < 38

7 38 to 50

Problems here are typically asso ciated with sphere heating while aw aiting the start of sweep

trigger. If necessary adjust the squegging clamp calibration constant for the appropriate band.

Single-Band Amplifier/Filter Delay .

95. On the syn thesizer, set:

2.0

ST ART

STOP

CONT

SWEEP TIME

POWER LEVEL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Leveling Mode ALCoff (asterisk on)

ALC

F or the 83640B /L and 83650B /L only , also set:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Disable Doubler (asterisk on)

SERVICE

96. Decrease the syn thesizer p ow er lev el un til the lev el of the en tire trace on the oscilloscop e

b egins to drop.

Show 34 more labels / short text on this page
Band 83630B /L · F requency Range · F requency Range · F requency Range · (GHz) · (GHz) · (GHz) · · · · GHz · · · 7 · GHz · · · · · 0 · msec · · · · 25 · dBm · · · · · · Tools Menu · Agilent 8360 Adjustments · Amplifier/Filter Adjustments
Page 311 of the Agilent Technologies 83640B user manual — STOP
Page 311 · click to enlarge
Page 311

The oscilloscop e v ertical sensitivit y is initially set to 50 mV/division (5

mV/division with a 10:1 prob e) providing a logarithmic scale of 1.5

dB/division. If necessary , decrease the v ertical sensitivity for some frequency

bands to 100 mV/division in order to display the en tire trace on the CR T

(the oscilloscop e is already set to 10 mV/division for the 83623B /L and the

83624B ). If y ou do c hange the v ertical sensitivit y b y half, remem b er that the

logarithmic scale m ust c hange b y half also.

Band 1 Adjustment.

97. Switch b et w een the fastest sweep time (0 ms) and 200 ms sweep time and c hec k that

p ow er c hanges less than 2/3 division ( 1 dB.) If not, adjust the appropriate rise or dela y

term (steps 98 through 100) and rep eat this step.

If fast sweep has more p ow er than slow sweep do not readjust. Con tinue with

the next step.

98. On the syn thesizer, set:

SWEEP TIME

99. Press

SINGLE

sev eral times to initiate sev eral sweeps and c hec k that a drop in p ow er 1

division o ccurs when in single sweep. If it is less than 1 division, skip the remainder of

\Band 1 Adjustmen t" and con tinue with \Band 2 Adjustmen t."

If the drop in p ow er is > 1 division, increase the A term of the dela y adjustmen t #561 to

put the A term at the high end of the p eak.

Then rep eat from step 95.

100. Adjust calibration constant #561, YTF Dly T erm A Hrm 1, to maximize p ow er for the

low end of band 1. Set the calibration constant for the middle of the p eak ed range.

101. Adjust calibration constant #571, YTF Dly T erm B Hrm 1, to maximize p ow er for the

high end of band 1. Set the calibration constant for the middle of the p eak ed range.

102. Adjust calibration constant #205, YTM Rise; Band 1, to minimize p ow er drop out at the

start of band 1.

Band 2 Adjustment.

If the oscilloscop e is not already set to 5 mV/division

(or 10 mV/division for the 83623B /L and the 83624B ), do so no w. The

logarithmic scale is calculated with the 5 mV/division v alue so for mo dels

using 10 mV/division the logarithmic scale m ust c hange b y half.

103. On the syn thesizer, set:

ST ART

13.5

STOP

CONT

SWEEP TIME

POWER LEVEL

104. Decrease the syn thesizer p ow er lev el un til the lev el of the en tire trace on the oscilloscop e

b egins to drop.

105. Switch b et w een the fastest sweep time (0 ms) and 200 ms sweep time and c hec k that

p ow er c hanges less than 2/3 division. If not, adjust the appropriate rise or delay term

(steps 106 through 108) and then rep eat this step.

Show 27 more labels / short text on this page
· · 0 · msec · · · · 7 · GHz · · · · · GHz · · · · 0 · · msec · · · 25 · · dBm · 3-52 Adjustments · Amplifier/Filter Adjustments
Page 312 of the Agilent Technologies 83640B user manual — SINGLE
Page 312 · click to enlarge
Page 312
106. On the syn thesizer, set:

SWEEP TIME

107. a. Press

sev eral times to initiate sev eral sweeps. Note an y drop in p ow er.

SINGLE

b. Press

SINGLE

RF ON/OFF

. Then press

RF ON/OFF

again (LED on) and note an y drop

in p ow er.

c. The p ow er loss noted in step a can b e no smaller than the p ow er loss noted in step b.

If the step a p ow er loss is greater than the step b p ow er loss b y 1 division ( 1.5 dB),

increase the A term of the dela y adjustmen t #562 to put the A term at the high end

of the p eak. Then rep eat from step 103.

If the p ow er loss in step a is not smaller than that in step b, skip the rest of \Band 2

Adjustmen t" and con tinue with \Band 3 Adjustmen t."

108. Adjust calibration constant #562, YTF Dly T erm A Hrm 2, to maximize p ow er for the

low end of band 2.

109. Adjust calibration constant #572, YTF Dly T erm B Hrm 2, to maximize p ow er for the

high end of band 2.

110. Adjust calibration constant #208, YTM Rise; Band 2 A2, to minimize p ow er drop out at

the start of band 2.

Band 3 Adjustment.

111. On the syn thesizer, set:

13.5

ST ART

STOP

CONT

SWEEP TIME

POWER LEVEL

112. Decrease the syn thesizer p ow er lev el un til the lev el of the en tire trace on the oscilloscop e

b egins to drop.

113. Switch b et w een the fastest sweep time (0 ms) and 200 ms sweep time and c hec k that

p ow er c hanges less than 2/3 division ( 1 dB). If not, adjust the appropriate rise or dela y

terms (steps 114 through 116) and then rep eat this step.

114. On the syn thesizer, set:

SWEEP TIME

115. a. Press

SINGLE

sev eral times to initiate sev eral sweeps. Note an y drop in p ow er.

b. Press

. Then press

again (LED on) and note an y drop

SINGLE

RF ON/OFF

RF ON/OFF

in p ow er.

c. The p ow er loss noted in step a can b e no smaller than the p ow er loss noted in step b.

If the step a p ow er loss is greater than the step b p ow er loss b y 1 division ( 1.5 dB),

increase the A term of the dela y adjustmen t #573 to put the A term at the high end

of the p eak. Then rep eat from step 111.

If the p ow er loss in step a is not smaller than that in step b, skip the rest of \Band 3

Adjustmen t" and con tinue with \Band 4 Adjustmen t."

116. Adjust calibration constant #563, YTF Dly T erm A Hrm 3, to maximize p ow er for the

low end of band 3.

117. Adjust calibration constant #573, YTF Dly T erm B Hrm 3, to maximize p ow er for the

high end of band 3.

118. Adjust calibration constant #213, YTM Rise; Band 3 A3, to minimize p ow er drop out at

the start of band 3.

Show 34 more labels / short text on this page
· · 0 · · msec · · · · · GHz · · · 20 · · GHz · · · · 0 · msec · · · · 25 · dBm · · · · 0 · msec · · · Agilent 8360 Adjustments · Amplifier/Filter Adjustments
Page 313 of the Agilent Technologies 83640B user manual — RF ON/OFF
Page 313 · click to enlarge
Page 313

Band 4 Adjustment (83630B/L, 83640B/L, and 83650B/L).

119. On the syn thesizer, set:

ST ART

26.5

( 83630B /L and 83650B /L )

STOP

STOP

25.5

( 83640B /L)

CONT

SWEEP TIME

POWER LEVEL

120. Decrease the syn thesizer p ow er lev el un til the lev el of the en tire trace on the oscilloscop e

b egins to drop.

121. Switch b et w een the fastest sweep time (0 ms) and 200 ms sweep time and c hec k that

p ow er c hanges less than 2/3 division. If not, adjust the appropriate rise term (steps

122 and 123) and then rep eat this step.

122. On the syn thesizer, set:

SWEEP TIME

123. a. Press

sev eral times to initiate sev eral sweeps. Note an y drop in p ow er.

SINGLE

b. Press

. Then press

again (LED on) and note an y drop

SINGLE

RF ON/OFF

RF ON/OFF

in p ow er.

c. The p ow er loss noted in step a can b e no smaller than the p ow er loss noted in step b.

If the step a p ow er loss is greater than the step b p ow er loss b y 1 division ( 1.5 dB),

increase the A term of the dela y adjustmen t #564 to put the A term at the high end

of the p eak. Then rep eat from step 119.

If the p ow er loss in step a is not smaller than that in step b, skip the rest of \Band 4

Adjustmen t" and con tinue with \Band 5 Adjustmen t."

124. Adjust calibration constant #564, YTF Dly T erm A Hrm 4, to maximize p ow er for the

low end of band 4.

125. Adjust calibration constant #574, YTF Dly T erm B Hrm 4, to maximize p ow er for the

high end of band 4.

Band 5 Adjustment (83640B/L and 83650B/L only).

126. On the syn thesizer, set:

ST ART

25.5

( 83640B /L)

ST ART

26.5

( 83650B /L)

STOP

( 83640B /L)

33.5

( 83650B /L)

STOP

CONT

SWEEP TIME

POWER LEVEL

127. Decrease the syn thesizer p ow er lev el un til the lev el of the en tire trace on the oscilloscop e

b egins to drop.

128. Switch b et w een the fastest sweep time (0 ms) and 200 ms sweep time and c hec k that

p ow er c hanges less than 2/3 division ( 1 dB). If not, adjust the appropriate rise or delay

term (steps 129 through 131) and then rep eat this step.

129. On the syn thesizer, set:

SWEEP TIME

130. a. Press

SINGLE

sev eral times to initiate sev eral sweeps. Note an y drop in p ow er.

b. Press

SINGLE

RF ON/OFF

. Then press

RF ON/OFF

again (LED on) and note an y drop

in p ow er.

c. The p ow er loss noted in step a can b e no smaller than the p ow er loss noted in step b.

If the step a p ow er loss is greater than the step b p ow er loss b y 1 division ( 1.5 dB),

increase the A term of the delay adjustmen t #565 to put the A term at the high end

of the p eak. Then rep eat from step 126.

Show 60 more labels / short text on this page
· · 20 · · GHz · · · GHz · · · GHz · · · · 0 · msec · · · · 25 · dBm · · · · 0 · msec · · · · · GHz · · · GHz · · · 32 · GHz · · · GHz · · · · 0 · · msec · · · 25 · · dBm · · · 0 · msec · · · 3-54 Adjustments · Amplifier/Filter Adjustments
Page 314 of the Agilent Technologies 83640B user manual — ST ART
Page 314 · click to enlarge
Page 314

If the p ow er loss in step a is not smaller than that in step b, skip the rest of \Band 5

Adjustmen t" and con tinue with \Band 6 Adjustmen t."

131. Adjust calibration constant #565, YTF Dly T erm A Hrm 5, to maximize p ow er for the

low end of band 5.

132. Adjust calibration constant #575, YTF Dly T erm B Hrm 5, to maximize p ow er for the

high end of band 5.

133. Adjust calibration constant #215, YTM Rise; Band 5, to minimize p ow er drop out at the

start of band 5.

Band 6 Adjustment (83640B/L and 83650B/L only).

134. On the syn thesizer, set:

ST ART

( 83640B /L)

ST ART

33.5

( 83650B /L)

STOP

( 83640B /L)

STOP

( 83650B /L)

CONT

SWEEP TIME

POWER LEVEL

135. Decrease the syn thesizer p ow er lev el un til the lev el of the en tire trace on the oscilloscop e

b egins to drop.

136. Switch b et w een the fastest sweep time (0 ms) and 200 ms sweep time and c hec k that

p ow er c hanges less than 2/3 division ( 1 dB). If not, adjust the appropriate rise or dela y

term (steps 137 through 139) and then rep eat this step.

137. On the syn thesizer, set:

SWEEP TIME

138. a. Press

SINGLE

sev eral times to initiate sev eral sweeps. Note an y drop in p ow er.

b. Press

SINGLE

RF ON/OFF

. Then press

RF ON/OFF

again (LED on) and note an y drop

in p ow er.

c. The p ow er loss noted in step a can b e no smaller than the p ow er loss noted in step b.

If the step a p ow er loss is greater than the step b p ow er loss b y 1 division ( 1.5 dB),

increase the A term of the dela y adjustmen t #566 to put the A term at the high end

of the p eak. Then rep eat from step 134.

If the p ow er loss in step a is not smaller than that in step b, skip the rest of \Band 6

Adjustmen t" and con tinue with \Band 7 Adjustmen t."

139. Adjust calibration constant #566, YTF Dly T erm A Hrm 6, to maximize p ow er for the

low end of band 6.

140. Adjust calibration constant #576, YTF Dly T erm B Hrm 6, to maximize p ow er for the

high end of band 6.

141. Adjust calibration constant #216, YTM Rise; Band 6, to minimize p ow er drop out at the

start of band 6.

Band 7 Adjustment (83650B/L only).

142. On the syn thesizer, set:

ST ART

STOP

CONT

SWEEP TIME

POWER LEVEL

143. Decrease the syn thesizer p ow er lev el un til the lev el of the en tire trace on the oscilloscop e

b egins to drop.

Show 55 more labels / short text on this page
· · 32 · GHz · · · GHz · · · 40 · GHz · · · 38 · GHz · · · · 0 · · msec · · · 25 · · dBm · · · 0 · · msec · · · · 38 · · GHz · · · 50 · GHz · · · · · 0 · msec · · · · 25 · dBm · · Agilent 8360 Adjustments · Amplifier/Filter Adjustments
Page 315 of the Agilent Technologies 83640B user manual — MARKER
Page 315 · click to enlarge
Page 315
144. Switch b et w een the fastest sweep time (0 ms) and 200 ms sweep time and c hec k that

p ow er c hanges less than 2/3 division ( 1 dB). If not, adjust the appropriate rise or dela y

term (steps 145 through 147) and then rep eat this step.

145. On the syn thesizer, set:

SWEEP TIME

146. a. Press

SINGLE

sev eral times to initiate sev eral sweeps. Note an y drop in p ow er.

b. Press

. Then press

again (LED on) and note an y drop

SINGLE

RF ON/OFF

RF ON/OFF

in p ow er.

c. The p ow er loss noted in step a can b e no smaller than the p ow er loss noted in step b.

If the step a p ow er loss is greater than the step b p ow er loss b y 1 division ( 1.5 dB),

increase the A term of the dela y adjustmen t #567 to put the A term at the high end

of the p eak. Then rep eat from step 142.

If the p ow er loss in step a is not smaller than that in step b, skip the rest of \Band 7

Adjustmen t" and con tinue with \Multi-band YTF Delay ."

147. Adjust calibration constant #567, YTF Dly T erm A Hrm 7, to maximize p ow er for the

low end of band 7.

148. Adjust calibration constant #577, YTF Dly T erm B Hrm 7, to maximize p ow er for the

high end of band 7.

149. Adjust calibration constant #217, YTM Rise; Band 7, to minimize p ow er drop out at the

start of band 7.

Multi-band YTF Delay .

150. On the syn thesizer, set:

ST ART

STOP

CONT

SWEEP TIME

POWER LEVEL

151. Decrease the syn thesizer p ow er lev el un til the lev el of the en tire trace on the oscilloscop e

b egins to drop.

Syn thesizer in tensity markers are set to iden tify the bandcross frequencies (7.0,

13.5 GHz, and 20.0 GHz). Reduce oscilloscop e trace in tensity to iden tify these

markers. Y ou can also iden tify bandcrosses b y temp orarily removing the cable

from the Z-AXIS BLANK/MKRS connector on the syn thesizer rear panel.

Make sure this cable is connected when making adjustmen ts.

152. On the syn thesizer, set:

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

13.5

153. Adjust calibration constant #593, YTF Bx Dly T erm A3, to maximize p ow er at the low

end of band 3.

154. Adjust calibration constant #603, YTF Bx DL Y T erm B3 to maximize p ow er at the high

end of band 3.

155. Adjust calibration constant #212, YTM Rise; Band 3 B2, to minimize p ow er drop out at

the end of band 3.

156. Switch b et w een the fastest sweep time (0 ms) and 200 ms sweep time and c hec k that

p ow er c hanges less than 2/3 division ( 1 dB). If the c hange in p ow er is greater than 2/3

division, readjust the A term of the delay adjustmen t #593 for low end problems, or the

B term of the dela y adjustmen t #603 for high end problems.

Show 34 more labels / short text on this page
· · 0 · msec · · · · · 13 · GHz · · · · 20 · GHz · · · · · 0 · · msec · · · 25 · · dBm · · · Marker M1 · GHz · (asterisk on) · 3-56 Adjustments · Amplifier/Filter Adjustments
Page 316 of the Agilent Technologies 83640B user manual — ST ART
Page 316 · click to enlarge
Page 316
157. On the syn thesizer, set:

CONT

SWEEP TIME

158. Press

SINGLE

sev eral times to initiate sev eral sweeps. If a drop in p ow er greater than 1

division o ccurs when in single sweep, increase the A term of the dela y adjustmen t #593

to put the A term at the high end of the p eak.

159. On the syn thesizer, set:

ST ART

CONT

SWEEP TIME

POWER LEVEL

160. Decrease the syn thesizer p ow er lev el un til the lev el of the en tire trace on the oscilloscop e

b egins to drop.

161. Switch b et w een the fastest sweep time (0 ms) and 500 ms sweep time and c hec k that

p ow er c hanges less than 2/3 division ( 1 dB) across b oth bands. If the c hange is greater

than 2/3 division, readjust the appropriate dela y term.

162. Adjust calibration constant #592, YTF Bx Dly T erm A2, to maximize p ow er at the start

of band 2 (start of sweep).

163. Adjust calibration constant #602, YTF Bx Dly T erm B2, to maximize p ow er at the end

of band 2.

164. Adjust calibration constant #211, YTM Rise; Band 3 A2, to minimize p ow er drop out at

the start of band 3.

165. On the syn thesizer, set:

ST ART

POWER LEVEL

166. Decrease the syn thesizer p ow er lev el un til the lev el of the en tire trace on the oscilloscop e

b egins to drop.

167. Adjust calibration constant #212, YTM Rise; Band 3 B2, to minimize p ow er drop out at

the start of band 3.

168. On the syn thesizer, set:

SWEEP TIME

169. Press

SINGLE

sev eral times to initiate sev eral sweeps. If a drop in p ow er greater than 1

division o ccurs when in single sweep, readjust the appropriate dela y term for the lo cation

of the problem.

170. On the syn thesizer, set:

2.0

ST ART

CONT

SWEEP TIME

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWER LEVEL

171. Decrease the syn thesizer p ow er lev el un til the lev el of the en tire trace on the oscilloscop e

b egins to drop.

172. Adjust calibration constant #591, YTF Bx Dly T erm A1, to maximize p ow er at the start

of band 1.

173. Adjust calibration constant #601, YTF Bx Dly T erm B1, to maximize p ow er at the high

end of band 1.

174. Adjust calibration constant #206, YTM Rise; Band 2 A1, to minimize p ow er drop out at

the start of band 2.

175. Adjust calibration constant #209, YTM Rise; Band 3 A1, to minimize p ow er drop out at

the start of band 3.

176. Switch b et w een the fastest sweep time (0 ms) and 1s sweep time and c hec k that p ow er

c hanges less than 2/3 division ( 1 dB) across all bands. If the c hange is greater than 2/3

division, readjust the appropriate delay term.

Show 60 more labels / short text on this page
· · · 0 · · msec · · · 7 · · GHz · · · · 0 · · msec · · · 25 · · dBm · · · 10 · GHz · · · · 25 · dBm · · · · 0 · msec · · · · · GHz · · · · 0 · msec · · · · Marker M2 · 7 · GHz · (asterisk on) · · · 25 · · dBm · Agilent 8360 Adjustments · Amplifier/Filter Adjustments
Page 317 of the Agilent Technologies 83640B user manual — SINGLE
Page 317 · click to enlarge
Page 317
177. On the syn thesizer, set:

SWEEP TIME

178. Press

SINGLE

sev eral times to initiate sev eral sweeps. If a drop in p ow er greater than 1

division o ccurs when in single sweep, readjust the appropriate dela y or risetime term for

the lo cation of the problem.

179. On the syn thesizer, set:

ST ART

POWER LEVEL

180. Decrease the syn thesizer p ow er lev el un til the lev el of the en tire trace on the oscilloscop e

b egins to drop.

181. Adjust calibration constant #210, YTM Rise; Band 3 B1, to minimize p ow er drop out at

the end of band 3.

182. Adjust calibration constant #207, YTM Rise; Band 2 B1, to minimize p ow er drop out at

the end of band 2.

Band 4 Multi-band YTF Delay (83630B/L, 83640B/L, and 83650B/L only).

183. On the syn thesizer, set:

ST ART

2.0

STOP

26.5

( 83630B /L and 83650B /L )

25.5

( 83640B /L)

STOP

CONT

SWEEP TIME

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER

POWER LEVEL

184. Decrease the syn thesizer p ow er lev el un til the lev el of the en tire trace on the oscilloscop e

b egins to drop.

185. Adjust calibration constant #594, YTF Bx Dly T erm A4, to maximize p ow er at the start

of band 4.

186. Adjust calibration constant #604, YTF Bx Dly T erm B4, to maximize p ow er at the end

of band 4.

187. Adjust calibration constant #214, YTM Rise; Band 4, to minimize p ow er drop out at the

start of band 4.

188. Switch b et w een the fastest sweep time (0 ms) and 500 ms sweep time and c hec k that

p ow er c hanges less than 2/3 division ( 1 dB) across b oth bands. If the c hange is greater

than 2/3 division, readjust the appropriate dela y term.

189. On the syn thesizer, set:

SWEEP TIME

190. Press

sev eral times to initiate sev eral sweeps. If a drop in p ow er greater than 1

SINGLE

division o ccurs when in single sweep, readjust the appropriate delay term for the lo cation

of the problem.

191. On the syn thesizer, set:

ST ART

CONT

SWEEP TIME

POWER LEVEL

192. Decrease the syn thesizer p ow er lev el un til the lev el of the en tire trace on the oscilloscop e

b egins to drop.

193. Press

SINGLE

sev eral times to initiate sev eral sweeps. If a drop in p ow er greater than 1/3

division o ccurs when in single sweep, readjust the appropriate delay or risetime term for

the lo cation of the problem.

Show 65 more labels / short text on this page
· · 0 · · msec · · · 5 · · GHz · · · 25 · · dBm · · · GHz · · · · GHz · · · GHz · · · · 0 · · msec · · · Marker M3 · 20 · GHz · (asterisk on) · · · 25 · · dBm · · · 0 · · msec · · · 15 · GHz · · · · · 0 · · msec · · · 25 · · dBm · 3-58 Adjustments · Amplifier/Filter Adjustments
Page 318 of the Agilent Technologies 83640B user manual — 83620B /
Page 318 · click to enlarge
Page 318

Operation Check.

194. T ry the following start and stop frequencies:

Start and Stop Frequencies (GHz)

83620B /

83622B /24B 83630B /L 83640B/L 83650B /L

83623B/L

0.045 to 20 2 to 20 0.045 to 26.5 0.045 to 40 0.045 to 50

4.5 to 18 4.5 to 18 4.5 to 18 4.5 to 32 4.5 to 33

10 to 20 10 to 20 10 to 20 10 to 23 10 to 45

2 to 14.5 2 to 14.5 2 to 14.5 2 to 23 2 to 23

Compare sweeps of 0 ms and 1 s and also single sweeps at fast sp eed. If a p ow er drop of

0.5 division or greater is noted, readjust the appropriate calibration constant.

195. If this is the last calibration constant y ou will b e adjusting, see \Calibration Constants"

to store the calibration constants as protected data (in EEPR OM).

Related P erformance T ests

Maximum Lev eled P ow er

In Case Of Difficulty

1. See \Adjustmen t Help" in this pro cedure.
2. V erify the oscilloscop e ALC b oard calibration. When calibrated, the v oltage on the

oscilloscop e is scaled to 33 mV/dB.

3. Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

Agilent 8360 Adjustments · Amplifier/Filter Adjustments

Page 319 of the Agilent Technologies 83640B user manual — CW
Page 319 · click to enlarge
Page 319
9. Low P o w er SRD Bias

The syn thesizer is set to 10 GHz. Then the A9 pulse b oard MIN adjustmen t sets the source

v oltage of the FET for the prop er SRD bias lev el.

1. Connect the equipmen t as sho wn in Figure 3-21 . Preset all instrumen ts and let them w arm

up for at least one hour.

Agilent Technologies 83640B — 08360125.pdf diagram (page 319)
Fig. 101 · 08360125.pdf

Figure 3-21. Lo w P o w er SRD Bias Adjustment Setup

2. On the syn thesizer set:

.

CW

3. Set the DVM to measure DC v olts.
4. Adjust A9R105 MIN (see Figure 3-22) for 0.54 V DC.

Lo w P o w er SRD Bias

Show 8 more labels / short text on this page
Description and Procedure · Equipment · DVM HP/Agilent 3456A · · · 10 · GHz · 3-60 Adjustments
Page 320 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 320 · click to enlarge
Page 320
Agilent Technologies 83640B — 08360125.pdf diagram (page 320)
Fig. 102 · 08360125.pdf

Figure 3-22. A9R105 Location

Related P erformance T ests

In Case of Difficulty

Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series Swept

CW Gener ator T r oublesho oting Guide .

Lo w P o w er SRD Bias

None · Agilent 8360 Adjustments

Page 321 of the Agilent Technologies 83640B user manual — ENTER
Page 321 · click to enlarge
Page 321
10. Modulator Offset and Gain

In this pro cedure, the ALC mo dulation o set and gain calibration constants are adjusted

to linearize the ALC mo dulator resp onse to the ALC p ow er lev el reference v oltage. Default

v alues are en tered for the mo dulator o set calibration constants and in ternal rm w are is

activ ated to set the mo dulator gain calibration constants. The syn thesizer is then set for a

p ow er sweep across the en tire lev eled ALC range ( 20 to the maximum lev eled p ow er). The

in tegrator lev el signal on the ALC b oard is monitored to v erify linearity .

If necessary , the mo dulator o set v alues are mo di ed.

1. F or 8360 B-Series only , connect the equipmen t as sho wn in Figure 3-23. Preset the

instrumen ts and let them w arm up for at least one hour. F or 8360 L-Series, start this test

at step 2.

Agilent Technologies 83640B — 08360125.pdf diagram (page 321)
Fig. 103 · 08360125.pdf

Figure 3-23. Modulator Offset and Gain Adjustment Setup

Digitizing Oscilloscop e HP/Agilent 54111D

Oscilloscop e Prob es HP/Agilent 10431A

2. On the syn thesizer, set:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SERVICE

3. If y ou ha v e replaced the low band assembly (A24) or the dual mo dulator assembly

(A38), set the ALC o set calibration constants to the default v alues given in T able 3-6 .

Otherwise, skip to step 7 for 8360 B-Series or skip to the next adjustmen t for 8360

L-Series.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Select Cal . Using the n umeric k eypad, en ter the n um b er of the rst calibration

constant in T able 3-6 . T erminate y our en try with the

ENTER

k ey .

Modulator Offset and Gain

Show 8 more labels / short text on this page
Description and Procedure · Equipment · · · Adjust Menu · Calib Menu · Select · 3-62 Adjustments
Page 322 of the Agilent Technologies 83640B user manual — ENTER
Page 322 · click to enlarge
Page 322

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Modify Cal . Use the n umeric k eypad to en ter the corresp onding default v alue in

Figure 3-23 . T erminate y our en try with the

ENTER

k ey .

4. Using the up/do wn arrow k eys, select eac h of the rest of the calibration constants in

T able 3-6 and set their default v alues b y en tering the v alue on the n umeric k eypad.

T erminate y our en try with the

ENTER

k ey .

T able 3-6. ALC Offset Calibration Constant Default V alues

Description Default V alue

315 ALC Mo d Ofs; Band 0 100

316 ALC Mo d Ofs; Band 1 100

317 ALC Mo d Ofs; Band 2 100

318 ALC Mo d Ofs; Band 3 100

319 ALC Mo d Ofs; Band 4 100

320 ALC Mo d Ofs; Band 5 150

321 ALC Mo d Ofs; Band 6 150

322 ALC Mo d Ofs; Band 7 150

Modulator Offset and Gain

Show 4 more labels / short text on this page
Select · Calibration · Constan t · Agilent 8360 Adjustments
Page 323 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 323 · click to enlarge
Page 323
5. On the syn thesizer, set:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SERVICE

The syn thesizer will measure, calculate, and store the mo dulator gain calibration

constants.

When the adjustmen t is completed the following message app ears on the display:

Modulator Gain Cal Completed

Calibration constants were modified.

6. If this is the last calibration constant y ou will b e adjusting, see \Calibration Constants"

to store the calibration constants as protected data (in EEPR OM).

8360 L-Series: This adjustmen t is complete; con tinue with next adjustmen t.

8360 B-Series: Con tinue with this adjustmen t.

7. Connect the oscilloscop e as sho wn in Figure 3-23.
8. On the oscilloscop e, set:

Channel 1:

Input Impedance 1 M

Channel 2:

Input Impedance 1 M

Timebase:

Delay Reference At left

Trigger:

Trigger Source Channel 2

Trigger Level 2 V

Display:

Number of Averages 2

Modulator Offset and Gain

Show 22 more labels / short text on this page
· · Adjust Menu · AssyAdj Menu · A10 Adj Menu · A10 Mod Gain · Display On · Volts/Division 50 mV · Offset As necessary · Input Coupling dc · Display On · Volts/Division 2 V · Input Coupling dc · Time/Division 10 ms · Delay 5 ms · Sweep Triggered · Trigger Mode Edge · Trigger Slope Negative · Display Mode Repetitive · Averaging On · Screen Single · 3-64 Adjustments
Page 324 of the Agilent Technologies 83640B user manual — USER CAL
Page 324 · click to enlarge
Page 324
9. On the syn thesizer, set:

CENTER

SP AN

10. F or syn thesizers with Option 001, set:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PO WER

MENU

11. Set the syn thesizer p ow er lev el to 20 dBm. Press:

POWER LEVEL

No w set the syn thesizer for a p ow er sweep of its full ALC range. F or example, if y our

syn thesizer's sp eci ed maxim um lev eled p ow er at this fr e quency is +10 dBm then set the

p ow er sweep to 30 dB/sw eep ( 20 dBm to +10 dBm = 30 dB). Press:

PO WER

MENU

POWER SWEEP

12. Con tinue to set up the syn thesizer as follo ws:

SWEEP TIME

USER CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Peak RF Always (asterisk on)

13. Adjust the oscilloscop e timebase so the Z-axis input go es high at the right edge of the

display . This signal is low during the p ow er sweep.

14. Change the v alue of the band 0 calibration constant #315 to adjust the signal v ariation

during the p ow er sweep (while Z-axis is low) as follo ws:

Adjust the calibration constant so that the trace in the last three (farthest right)

graticules is as at as p ossible. Initiate an AM bandwidth calibration, press:

USER CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM BW Cal Once .

Repeat the adjustmen t and calibration un til the last three graticules are as at as

p ossible.

15. Set the syn thesizer to eac h of the cen ter frequencies in T able 3-7 . Change the p ow er sweep

as appropriate to sweep the full ALC range. P erform an AM bandwidth calibration and

then rep eat step 14 for eac h corresp onding calibration constant.

T able 3-7. Center Frequencies and Calibration Constants

Cen ter F requencies

5.0 #316 Band 1

9.0 #317 Band 2

19.0 #318 Band 3

1 Disregard if b ey ond your synthesizer's capability .

Modulator Offset and Gain

Show 41 more labels / short text on this page
· · 1 · · GHz · · · 0 · · GHz · · Uncoupl Atten · · · · 20 · · · dBm · · · 30 · · dB · · · 100 · · msec · · · Tracking Menu · · AM Cal Menu · Calibration Constan ts · (GHz) · #319 Band 4 · #320 Band 5 · #321 Band 6 · #322 Band 7 · Agilent 8360 Adjustments
Page 325 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 325 · click to enlarge
Page 325

Related P erformance T ests

Maximum Lev eled P ow er AM Bandwidth

In Case of Difficulty

Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series Swept

CW Gener ator T r oublesho oting Guide .

Modulator Offset and Gain

3-66 Adjustments

Page 326 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 326 · click to enlarge
Page 326
11. ALC P o w er Lev el Accuracy

This adjustmen t sets the absolute p ow er accuracy in eac h frequency band at a CW frequency .

The digital ALC calibration arra y is loaded with zeros for the 0 dB attenuator setting to

eliminate an y p ow er o set from this source. Calibration constants set the p ow er accuracy at

four p ow er lev els.

A p oten tiometer sets the p ow er accuracy at an 18 dBm lev el in the low band

(not applicable for all mo dels).

This adjustmen t pro cedure zeros the digital ALC calibration arra y at the 0 dB

attenuator setting. The \Pow er Flatness" adjustmen t m ust b e p erformed after

this pro cedure.

1. Preset the equipmen t sho wn in Figure 3-24 and let them w arm up for at least one hour.

Do not connect the p ow er sensor to the syn thesizer RF OUTPUT y et.

Agilent Technologies 83640B — 08360125.pdf diagram (page 326)
Fig. 104 · 08360125.pdf

Figure 3-24. ALC P o w er Lev el Accuracy Adjustment Setup

P ow er Meter HP/Agilent 436A/7A/8A

P ow er Sensor HP/Agilent 8485A

(83620B/22B/24B, 83623B/L, and 83630B/L)

P ow er Sensor HP/Agilent 8487A (83640B/L and 83650B/L)

2. Zero and calibrate the p ow er meter/sensor and set the p ow er meter to dBm mo de.

Connect the p ow er sensor to the syn thesizer's RF OUTPUT.

Set the CAL F A CTOR % for the p ow er sensor 1 GHz calibration factor.

3. On the syn thesizer, set:

USER CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RF Peak Always (asterisk on)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PO WER

MENU

Uncoupl Atten (Option 001 only)

ALC P o w er Lev el Accuracy

Show 7 more labels / short text on this page
Description and Procedure · Equipment · · · Tracking Menu · · Agilent 8360 Adjustments
Page 327 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 327 · click to enlarge
Page 327
4. On the syn thesizer, zero the digital ALC calibration for the 0 dB attenuator setting for all

frequency ranges:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SERVICE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AtnStep To Cal

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Cal Freq Range Lo Band

ALC Cal Array (Atn, Freq) Zero

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Cal Freq Range Hi Band

ALC Cal Array (Atn, Freq) Zero

The default v alues given in the next step are a general starting p oint. If the

curren t ALC p ow er accuracy is within 1 dB, few er rep etitions are required if

y ou start with the existing calibration constants unc hanged.

5. Set the default v alues given in T able 3-8 in to the corresp onding ALC calibration

constants. On the syn thesizer, set:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SERVICE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Select Cal En ter the n um b er of the calibration constant from T able 3-8 and terminate

the en try with

.

ENTER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Modify Cal En ter the default v alue from T able 3-8 and terminate the en try with

ENTER

.

T able 3-8. ALC Calibration Constants and Default V alues

ALC Calibration Constan t Default V alue

Low Band 0.01 to 2.0 GHz

#265 L VL DA C Ofs Lo Bnd 120

#252 L VL DA C Gain Lo Bnd 20

#284 ALC Det Ofs; Lo 142

#293 ALC Log Brkpt; Lo 140

High Band 2.0 to 20 GHz

#264 L VL DA C Ofs Hi Bnd 0

#251 L VL DA C Gain Hi Bnd 20

#283 ALC Det Ofs; Hi 134

#292 ALC Log Brkpt; Hi 131

1 Disregard if b eyond your synthesizer's capability .

Lo w Band ALC P o w er Accuracy Adjustment

6. On the syn thesizer, set:

CW

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SERVICE

7. On the syn thesizer, set:

First p ow er lev el in T able 3-9

POWER LEVEL

8. Select the rst calibration constant in T able 3-9 :

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ENTER

ALC P o w er Lev el Accuracy

Show 25 more labels / short text on this page
· · Adjust Menu · DigALC Menu · 0 · · dB · · · Adjust Menu · Calib Menu · · · 1 · GHz · · · · Adjust Menu · Calib Menu · · Select Cal · 265 · · 3-68 Adjustments
Page 328 of the Agilent Technologies 83640B user manual — CW
Page 328 · click to enlarge
Page 328

On the syn thesizer, select

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Modify Cal and, using the rotary knob, mo dify the calibration

constant so that the p ow er meter and the p ow er lev el setting are the same.

10. Repeat steps 8 through 10 for eac h p ow er lev el and calibration constant given in

T able 3-9.

T able 3-9. P o w er Lev el and Calibration Constant Adjustment

P ow er Lev el

Calibration Constan t Adjustmen t

0.0 #265 L VL DA C Ofs Lo Bnd

10.0 #252 L VL DA C Gain Lo Bnd

20.0 #284 ALC Det Ofs; Lo

+10.0 #293 ALC Det Log Brkpt; Lo

11. On the syn thesizer, set:

POWER LEVEL

Adjust A10R125, +20 dBm, un til the p ow er meter and the p ow er lev el setting are the

same. Figure 3-25 sho ws the lo cation of A10R125.

Agilent Technologies 83640B — POWER LEVEL diagram (page 328)
Fig. 105

Figure 3-25. A10R125 Location

12. Repeat steps 8 through 12 un til the p ow er accuracy for eac h calibration constant is within

0.01 dBm.

High Band P o w er Accuracy Adjustment

13. On the syn thesizer, set:

CW

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SERVICE

ALC P o w er Lev el Accuracy

Show 16 more labels / short text on this page
(dBm) · · · 18 · · dBm · · · 10 · GHz · · · · Adjust Menu · Calib Menu · Agilent 8360 Adjustments
Page 329 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 329 · click to enlarge
Page 329
14. On the p ow er meter, set the CAL F A CTOR % for the p ow er sensor 10 GHz calibration

factor.

15. On the syn thesizer, set:

POWER LEVEL

First p ow er lev el in T able 3-10

16. Select the rst calibration constant in T able 3-10:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ENTER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

On the syn thesizer, select

Modify Cal and, using the rotary knob, mo dify the calibration

constant so that the p ow er meter and the p ow er lev el setting are the same.

18. Repeat steps 16 through 18 for eac h p ow er lev el and calibration constant given in

T able 3-10.

19. Repeat steps 16 through 19 un til the p ow er accuracy for eac h calibration constant is

within 0.01 dBm.

T able 3-10. P o w er Lev el and Calibration Constant Adjustment

P ow er Lev el

Calibration Constan t Adjustmen t

0.0 #264 L VL DA C Ofs Hi Bnd

10.0 #251 L VL DA C Gain Hi Bnd

20.0 #283 ALC Det Ofs; Hi

+10.0 #292 ALC Det Log Brkpt; Hi

20. If this is the last calibration constant y ou will b e adjusting, see \Calibration Constants" in

this man ual to store the calibration constants as protected data (in EEPR OM).

Related P erformance T ests

In Case of Difficulty

Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series Swept

CW Gener ator T r oublesho oting Guide .

ALC P o w er Lev el Accuracy

Show 7 more labels / short text on this page
· Select Cal · 264 · · (dBm) · None · 3-70 Adjustments
Page 330 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 330 · click to enlarge
Page 330
12. P o w er Flatness

(The automated p ow er atness adjustmen t in Chapter 5 may b e used in place of this

pro cedure.)

In this pro cedure, the syn thesizer measures and corrects p ow er atness. The syn thesizer

con trols the p ow er meter via GPIB while the p ow er meter is measuring the RF output.

F or eac h syn thesizer frequency band and a 0 dB attenuator setting, con trol is given to the

syn thesizer to measure and correct p ow er atness.

This adjustmen t requires an HP/Agilent 437B P ow er Meter. The correct

p ow er sensor calibration factors m ust b e loaded and selected. This pro cedure

cannot b e run with a con troller on the GPIB, nor can it b e run from a front

panel em ulator.

1. Preset the instrumen ts sho wn in Figure 3-26. Do not connect the p ow er sensor to the

syn thesizer RF output y et. Let b oth instrumen ts w arm up for at least one hour.

Agilent Technologies 83640B — 08360125.pdf diagram (page 330)
Fig. 106 · 08360125.pdf

Figure 3-26. P o w er Flatness Adjustment Setup

P ow er Meter HP/Agilent 437A

P ow er Sensor HP/Agilent 8482A

P ow er Sensor HP/Agilent 8485A (83620B /22B/24B,

83623B /L, and 83630B /L)

P ow er Sensor HP/Agilent 8487A (83640B /L and 83650B/L)

Lo w Band P o w er Flatness

Disr e gar d if b eyond your synthesizer's c ap ability.

2. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SERVICE

P o w er Flatness

Show 7 more labels / short text on this page
Description and Procedure · Equipment · · · Adjust Menu · DigALC Menu · Agilent 8360 Adjustments
Page 331 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 331 · click to enlarge
Page 331

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Cal Freq Range Lo Bnd (asterisk on)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AtnStep to Cal

3. Load and select the calibration factors for the correct p ow er sensor for this frequency

band in to the p ow er meter. Refer to the p ow er meter man ual for instructions.

Zero and calibrate the p ow er meter then connect the p ow er sensor to the syn thesizer RF

output.

On the syn thesizer, press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Pwr Mtr ALC Cal .

The syn thesizer measures p ow er atness and calculates a new atness array for the

frequency band and attenuator setting.

On the syn thesizer, press

NNNNNNNNNNNNNN

SAVE and con tinue to the next step.

High Band P o w er Flatness

6. Load and select the calibration factors for the correct p ow er sensor for this frequency

band in to the p ow er meter. Refer to the p ow er meter man ual for instructions.

Zero and calibrate the p ow er meter then connect the p ow er sensor to the syn thesizer RF

output.

7. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Cal Freq Range Hi Band (asterisk on)

8. Repeat steps 3 through 5.

Millimeter Band P o w er Flatness

Disr e gar d if b eyond your synthesizer's c ap ability.

9. Load and select the calibration factors for the correct p ow er sensor for this frequency

band in to the p ow er meter. Refer to the p ow er meter man ual for instructions.

Zero and calibrate the p ow er meter then connect the p ow er sensor to the syn thesizer RF

output.

10. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Cal Freq Range mm Band (asterisk on)

11. Repeat steps 3 through 5.

Related P erformance T ests

P ow er Flatness

In Case of Difficulty

1. The HP/Agilen t 437B P ow er Meter m ust b e used. Correct calibration factors m ust b e

loaded and selected.

2. Make sure that the only GPIB connection is b et w een the syn thesizer and the p ow er meter.

No con troller is allo w ed on the bus.

3. Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

P o w er Flatness

Show 4 more labels / short text on this page
0 · dB · · 3-72 Adjustments
Page 332 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNN
Page 332 · click to enlarge
Page 332
13. AM/FM D A C Offset and Gain (Option 002)

No test equipmen t is required for this pro cedure. This pro cedure calibrates the mo dulation

generator assembly b y adjusting the AM and FM DA Cs. The in ternal DVM measures the

DA C output v oltages. The DA C o set is then zero ed with this adjustmen t pro cedure and the

DA C gain is automatically calibrated with calibration constants.

1. T urn the syn thesizer on and press

. Let the syn thesizer w arm up for at least one

PRESET

hour.

2. Select the A8 adjustmen t men u. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SERVICE

NNNNNNNNNNNNNNNNNNNN

AM DAC adjustmen t is automatically selected (asterisk on).

4. Adjust A8R111 (see Figure 3-27 ) to cen ter the \needle" on the display .
Agilent Technologies 83640B — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 332)
Fig. 107

Figure 3-27. A8R111 and A8R105 Adjustment Locations

NNNNNNNNNNNNNNNNNNNN

FM DAC (asterisk on).

6. Adjust A9R105 (see Figure 3-27 ) to cen ter the \needle" on the display .

NNNNNNNNNNNNNN

Done .

The syn thesizer will display the following message:

8. If this is the last calibration constant y ou will b e adjusting, see \Calibration Constants" to

store the calibration constants as protected data

(in EEPR OM).

AM/FM D A C Offset and Gain

Show 12 more labels / short text on this page
(8360 B-Series Only) · Description and Procedure · · · Adjust Menu · AssyAdj Menu · A8 Adj Menu · The · Select · Select · Updating Cal Constants · Agilent 8360 Adjustments
Page 333 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 333 · click to enlarge
Page 333

Related P erformance T ests

In Case of Difficulty

Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

AM/FM D A C Offset and Gain

AM Accuracy · FM Accuracy · 3-74 Adjustments

Page 334 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 334 · click to enlarge
Page 334
14. AM Accuracy

This pro cedure consists of three adjustmen ts. Calibration constant #277, AM O set, is

adjusted for no c hange in RF output p ow er when AM is enabled but no v oltage is applied.

EXP AM CAL is adjusted for a 10 dB c hange in p ow er when 1.00 V is applied and AM

10 dB/V is selected. LIN AM CAL is adjusted so that a 0.80 V AM input (100%/V AM

selected) causes the RF output to decrease to 4.00% of its initial v alue.

1. T urn on the instrumen ts sho wn in Figure 3-28 and let them w arm up for at least one

hour.

2. Zero and calibrate the p ow er meter/sensor.
3. Connect the equipmen t as sho wn in Figure 3-28.
Agilent Technologies 83640B — 08360125.pdf diagram (page 334)
Fig. 108 · 08360125.pdf

Figure 3-28. AM Accuracy Adjustment Setup

Digital V oltmeter HP/Agielnt 3456A

F unction Generator HP/Agilent 3325A

P ow er Meter HP/Agilent 436A/7A/8A

P ow er Sensor HP/Agilent 8485A (83620B /22B/24B,

83623B /L, and 83630B /L)

P ow er Sensor HP/Agilent 8487A (83640B /L and 83650B/L)

Show 5 more labels / short text on this page
(8360 B-Series Only) · Description and Procedure · Equipment · Agilent 8360 Adjustments · AM Accuracy
Page 335 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 335 · click to enlarge
Page 335
4. Set the function generator for a 0 V DC rear panel output (the instrumen t front panel

output is connected to the syn thesizer's AM input.)

5. On the syn thesizer, press:

CW

POWER LEVEL

6. Set the p ow er meter to read relative p ow er (dB).
7. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM On/Off 100%/V (asterisk on)

MOD

or, for syn thesizers with Option 002, instead press:

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MOD

8. Adjust calibration constant #277, AM O set, for a 0 dB reading on the p ow er meter.

Press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SERVICE

ENTER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Modify Cal Use the rotary knob to adjust the calibration constant for a 0 dB reading on

the p ow er meter.

9. Set the p ow er meter to measure absolute log p ow er (dBm).
10. On the syn thesizer, press:

POWER LEVEL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MOD

AM On/Off 100%/V (asterisk o )

or, for syn thesizers with Option 002, instead press:

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM On/Off Ext (asterisk o )

MOD

11. Set the function generator to the front panel output.
12. Set the p ow er meter to read relative p ow er (dB).
13. Set the function generator to 1.0 V DC o set. V erify the correct v oltage on the DVM.
14. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM On/Off 10dB/V (asterisk on)

or, for syn thesizers with Option 002, instead press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM On/Off Ext (asterisk on)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM Type 10dB/V (asterisk on)

15. Adjust A10R35, EXP AM CAL, for a 10 dB p ow er meter reading. See Figure 3-29 for

the lo cation of A10R35.

Show 39 more labels / short text on this page
· · 5 · GHz · · · · 15 · · · dBm · AM Offset · · · · · AM Menu · AM On/Off Ext · AM Type 100%/V · · · Adjust Menu · Calib Menu · Select Cal · 277 · · Log AM Accuracy · · · 0 · · dBm · · · · · AM Menu · 3-76 Adjustments · AM Accuracy
Page 336 of the Agilent Technologies 83640B user manual — MOD
Page 336 · click to enlarge
Page 336
Agilent Technologies 83640B — 08360125.pdf diagram (page 336)
Fig. 109 · 08360125.pdf

Figure 3-29. A10R35 and A10R39 Locations

16. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM On/Off 10dB/V (asterisk o )

or, for syn thesizers with Option 002, instead press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM On/Off Ext (asterisk o )

17. On the syn thesizer, press:

POWER LEVEL

18. Set the p ow er meter to measure linear mo de (watts).

F or p ow er meters without a relativ e linear mo de, adjust the syn thesizer p ow er lev el for a

3.16 mW p ow er meter reading.

F or p ow er meters with relativ e linear mo de (%), set the p ow er meter to relativ e.

19. Set the function generator for a 0.8 V DC o set. V erify the correct v oltage on the DVM.
20. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MOD

AM On/Off 100%/V (asterisk on)

or, for syn thesizers with Option 002, instead press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM On/Off Ext (asterisk on)

MOD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM Type 100%/V (asterisk on)

21. Adjust A10R39, LIN AM CAL, for a 0.126 mW (4.00% relative) reading. See Figure 3-29

for the lo cation of A10R39.

Related P erformance T ests

In Case of Difficulty

Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

Show 13 more labels / short text on this page
Linear AM Accuracy · · · 5 · · dBm · · · · · AM Accuracy · Agilent 8360 Adjustments · AM Accuracy
Page 337 of the Agilent Technologies 83640B user manual — MOD
Page 337 · click to enlarge
Page 337
15. AM Offset

This adjustmen t applies to L mo dels with the pre xes listed b elow. This pro cedure uses

calibration constant n um b er 277 to adjust the maximum p ow er out in scalar mo de to match

the RF output amplitude with scalar o .

Mo del Pre x

83623L 3722A and ab ov e

83630L 3722A and ab ov e

83640L 3722A and ab ov e

83650L 3722A and ab ov e

Oscilloscop e HP/Agilent 1740 or HP/Agilent 54600B

Crystal Detector HP/Agilent 3330D (Negative Detector)

1. Connect the equipmen t as sho wn in Figure 3-30 and let them w arm up for at least one

hour.

2. Set the syn thesizer to the following settings:

POWER LEVEL

CW

NNNNNNNNNNNNNNNNNNNN

Scalar (no * indicates Scalar is o )

3. Set the oscilloscop e to:

DC

.5V/div or less

adjust o set for trace on screen

4. On the syn thesizer press:

MOD

NNNNNNNNNNNNNNNNNNNN

Scalar (* indicates Scalar is on)

Show 28 more labels / short text on this page
(8360 L-Series Only) · Description and Procedure · Equipmen t Required · · · 0 · dBm · · · · 5 · GHz · · · · Mo d · · · · mo de · · scale · · o set · · · 3-78 Adjustments · AM Offset
Page 338 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 338 · click to enlarge
Page 338
On the syn thesizer press

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Adjust Menu . En ter password if requested

SERVICE

(factory password 8360) and then press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Modify Cal using the RPG, arrow k eys, or n umeric k ey pad, c hange the v alue

of cal constant 277 so the maximum p ow er out with scalar on is equal to the pulse p ow er

out with scalar o .

6. If this is the last adjustmen t required, sav e the cal data b y pressing

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNN

yes .

Agilent Technologies 83640B — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 338)
Fig. 110

Figure 3-30. AM Offset Adjustment Setup

Show 15 more labels / short text on this page
· more 1/3 · Adjust Menu · Calib Menu · Select Cal · 2 · 7 · 7 · · Hz sec Enter · · Cal Utility Menu · Save Cal · Agilent 8360 Adjustments · AM Offset
Page 339 of the Agilent Technologies 83640B user manual — POWER LEVEL
Page 339 · click to enlarge
Page 339
16. AM Delay

In the ALC lo op, the in tegrator input from the ALC detector is dela y ed (phase shifted from

the reference input) through the ALC mo dulator, RF path, and ALC detector. Because of

this, amplitude mo dulation can p eak the AM frequency resp onse. In this adjustmen t, y ou set

the dela y in the reference input equal to the dela y in the input from the ALC detector.

The syn thesizer is set to maximum lev eled p ow er at a CW frequency , with amplitude

mo dulation at the maximum rate. Monitoring the INT L VL signal (the in tegrator output),

adjust the AM DELA Y to minimize an y A C v oltage presen t.

1. Preset the instrumen ts sho wn in Figure 3-31 and let them w arm up for at least one hour.
Agilent Technologies 83640B — 08360125.pdf diagram (page 339)
Fig. 111 · 08360125.pdf

Figure 3-31. AM Delay Adjustment Setup

Digital V oltmeter HP/Agilent 3456A

F unction Generator HP/Agilent 3325A

2. T o ac hiev e p eak p ow er, turn on RF p eaking. Press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Peak RF Always (asterisk on)

USER CAL

3. Connect the equipmen t as sho wn in Figure 3-31 .
4. On the syn thesizer, press:

CW

Set to 3 dB b elow maximum sp eci ed lev eled p ow er.

POWER LEVEL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MOD

AM On/Off 100%/V (asterisk on)

Show 16 more labels / short text on this page
(8360 B-Series Only) · Description and Procedure · Equipment · · · Tracking Menu · · · 10 · · GHz · · · · 3-80 Adjustments · AM Delay
Page 340 of the Agilent Technologies 83640B user manual — MOD
Page 340 · click to enlarge
Page 340

or, for syn thesizers with Option 002, instead press:

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM On/Off Ext (asterisk on)

MOD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AM Type 100%/V (asterisk on)

The function generator should ha v e a 50 output for an accurate output.

If necessary , use a BNC tee and monitor the output lev el while the function

generator is connected to the AM input.

5. On the function generator, press:
FunctionSinewave
Frequency100 kHz
DC Offset0 V
OutputFront Panel
Amplitude600 mV p-p
6. Adjust A10R55 (see Figure 3-32 ) for a minimum A C v oltage reading on the DVM.

Y ou should exp ect to see v ery small c hanges in v oltage.

Agilent Technologies 83640B — MOD diagram (page 340)
Fig. 112

Figure 3-32. AM Delay Adjustment Location

Related P erformance T ests

In Case of Difficulty

Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

Show 6 more labels / short text on this page
· · AM Menu · AM Bandwidth · Agilent 8360 Adjustments · AM Delay
Page 341 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 341 · click to enlarge
Page 341
17. FM Gain

This adjustmen t sets the gain on the FM driver assembly to match the YO FM coil

sensitivit y , and to provide a 10 MHz/V FM input sensitivit y . The function generator is set to

provide FM at a 1 MHz rate and a 2.404 MHz deviation. This corresp onds to a mo dulation

index of 2.404, whic h should result in no p ow er in the carrier. FM gain is adjusted to

minimize the RF output p ow er.

1. Connect the equipmen t as sho wn in Figure 3-33 . Preset all instrumen ts and let them w arm

up for at least one hour.

Agilent Technologies 83640B — 08360125.pdf diagram (page 341)
Fig. 113 · 08360125.pdf

Figure 3-33. FM Gain Adjustment Setup

F unction Generator HP/Agilent 3325A

Sp ectrum Analyzer HP/Agilent 8566B

The function generator sp eci ed is accurate enough to set the amplitude

without using a DVM. If y ou use a di eren t function generator, use a wide

bandwidth DVM, suc h as an HP/Agilent 3458A, to accurately set the v oltage.

2. On the function generator, press:
FunctionSinewave
Show 5 more labels / short text on this page
(8360 B-Series Only) · Description and Procedure · Equipment · 3-82 Adjustments · FM Gain
Page 342 of the Agilent Technologies 83640B user manual — MOD
Page 342 · click to enlarge
Page 342
Frequency1 MHz
DC Offset0 V
Amplitude480.8 mV p-p (169.96 mV rms)
3. On the sp ectrum analyzer, press:
Reference Level0 dBm
AttenuatorAuto
Center Frequency5 GHz
Frequency Span50 kHz
Resolution BW3 kHz
Video BW1 kHz
Sweep TimeAuto
SweepContinuous
4. On the syn thesizer, press:

CW

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MOD

FM On/Off 100kHz (asterisk on)

or, for syn thesizers with Option 002, instead press:

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FM On/Off Ext (asterisk on)

MOD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FM Coupling 100kHz (asterisk on)

5. Adjust A11R55 (see Figure 3-34 ) for a minimum signal on the sp ectrum analyzer.
Agilent Technologies 83640B — MOD diagram (page 342)
Fig. 114

Figure 3-34. FM Gain Adjustment Location

Related P erformance T ests

In Case of Difficulty

1. V erify the function generator output (amplitude and frequency) with an oscilloscop e. Use a

BNC tee with the function generator connected to the FM input.

2. Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

Show 14 more labels / short text on this page
· · 5 · · GHz · · · · · FM Menu · Self-tests · FM Accuracy · Agilent 8360 Adjustments · FM Gain
Page 343 of the Agilent Technologies 83640B user manual — MOD
Page 343 · click to enlarge
Page 343
18. Square W a v e Symmetry

Use this adjustmen t to set the syn thesizer mo dulation (27.78 kHz square w av e) for equal RF

on/o p erio ds. This is accomplished b y minimizing the ev en harmonics of the 27.78 kHz

square w av e.

1. Connect the equipmen t as sho wn in Figure 3-35. Preset b oth instrumen ts and let them

w arm up for at least one hour.

Figure 3-35. Square W a v e Symmetry Adjustment Setup

Sp ectrum Analyzer HP/Agilent 8566B

2. On the syn thesizer, press:

CW

POWER LEVEL

3. On the sp ectrum analyzer, press:
Center Frequency5 GHz
Frequency Span200 kHz

Make sure the signal is cen tered on the display .

4. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MOD

Pulse On/Off Scalar (asterisk on)

or, for syn thesizers with Option 002, instead press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Pulse On/Off Scalar (asterisk on)

MOD

5. Set the sp ectrum analyzer marker to the p eak of either 2nd sideband

(56 kHz aw a y).

6. Press the marker to cen ter frequency k ey (adjusts the 2nd sideband to the cen ter of the

display).

7. On the sp ectrum analyzer, press:
Frequency Span0 Hz

Square W a v e Symmetry

Show 19 more labels / short text on this page
Description and Procedure · High Band Adjustment · Equipment · · · 5 · · GHz · · · 5 · · dBm · · · · · Pulse Menu · 3-84 Adjustments
Page 344 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 344 · click to enlarge
Page 344
8. Adjust calibration constant #330, Slow Sym High Band, for a minimum reading on the

sp ectrum analyzer. Press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SERVICE

Use the rotary knob or arrow k eys to select calibration constant #330. Press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

and use the rotary knob to adjust the calibration constant.

Lo w Band Adjustment

Disr e gar d if b eyond your synthesizer's c ap ability.

9. Reset the sp ectrum analyzer frequency span to 200 kHz.
10. Repeat this adjustmen t pro cedure at a frequency of 1 GHz and adjust calibration constant

#329 (Slow Sym Lo w Band) for a minimum reading on the sp ectrum analyzer.

11. If this is the last calibration constant y ou will b e adjusting, see \Calibration Constants"

to store the calibration constants as protected data (in EEPR OM).

Related P erformance T ests

In Case of Difficulty

Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

Square W a v e Symmetry

Show 8 more labels / short text on this page
· · Adjust Menu · Calib Menu · Select Cal · Modify Cal · None · Agilent 8360 Adjustments
Page 345 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 345 · click to enlarge
Page 345
19. AM Input Impedance

No test equipmen t is required for this pro cedure.

This pro cedure sets the AM input imp edance to either 50 or 2 k . The AM input

imp edance on A10 m ust b e set to 2 k for prop er op eration of the in ternal mo dulation

generator (Option 002).

Instrumen ts without Option 002 should p erform step A, \A10 Jump er".

Instrumen ts with Option 002 should set the A10 jump er to the 2 k p osition using step A,

\A10 Jump er" and should p erform step B, \A8 Jump er".

Show 4 more labels / short text on this page
(8360 B-Series Only) · Description and Procedure · 3-86 Adjustments · AM Input Impedance
Page 346 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 346 · click to enlarge
Page 346

A. A10 Jumper

1. On the syn thesizer, set the line p ow er switch to standb y .
2. Lift up the A10 ALC assembly far enough to access the jump er sho wn in Figure 3-36.
3. The AM input imp edance is factory-set to the 50 ohm p osition. T o c hange it to 2 k , mov e

the jump er to the p osition marked \2K". The jump er can b e returned to the \50 OHM"

p osition to c hange the AM input imp edance to 50 ohms.

4. Reinstall the A10 assembly .
Agilent Technologies 83640B — 08360125.pdf diagram (page 346)
Fig. 115 · 08360125.pdf

Figure 3-36. A10 Jumper Locations

Agilent 8360 Adjustments · AM Input Impedance

Page 347 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 347 · click to enlarge
Page 347

B. A8 Jumper

1. On the syn thesizer, make sure the line p ow er switch is set to standb y .
2. Lift up the A8 mo dulation generator assembly far enough to access the jump er sho wn in

Figure 3-37.

3. The AM input imp edance is factory-set to the 50 ohm p osition. T o c hange it to 2 k ,

mov e the jump er to the p osition marked \2K". The jump er can b e returned to the p osition

marked \50" to c hange the AM input imp edance to 50 ohms.

4. Reinstall the A8 assembly .
Agilent Technologies 83640B — 08360125.pdf diagram (page 347)
Fig. 116 · 08360125.pdf

Figure 3-37. A8 Jumper Locations

Related P erformance T ests

In Case of Difficulty

If the jump er is missing, see \Replaceable P arts".

Show 5 more labels / short text on this page
AM Accuracy · AM Bandwidth · AM Dynamic Range · 3-88 Adjustments · AM Input Impedance
Page 348 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 348 · click to enlarge
Page 348
20. FM Input Impedance

A soldering iron is required for this pro cedure.

This pro cedure sets the FM input imp edance to either 50 or 600 . The FM input

imp edance on A11 m ust b e set to 600 for prop er op eration of the in ternal mo dulation

generator (Option 002).

Instrumen ts without Option 002 should p erform step A, \A11 Jump er".

Instrumen ts with Option 002 should set the A11 jump er to the 600 p osition using step A,

\A11 Jump er" and should p erform step B, \A8 Jump er".

Show 4 more labels / short text on this page
(8360 B-Series Only) · Description and Procedure · Agilent 8360 Adjustments · FM Input Impedance
Page 349 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 349 · click to enlarge
Page 349

A. A11 Jumper

1. On the syn thesizer, set the line p ow er switch to standb y .
2. Disconnect the coaxial cables from the A11 FM driver assembly .
3. Remo v e the A11 assembly from the syn thesizer.
4. Lo cate the zero ohm resistor sho wn in Figure 3-38. The FM input imp edance is factory-set

to the 50 ohm p osition. T o c hange it to 600 ohms, remov e the resistor from the A11

assembly b y unsoldering it. The FM input imp edance can b e reset to 50 ohms b y soldering

the resistor bac k in to the A11 assembly .

If y ou will nev er use the 50 ohm setting, the resistor can b e clipp ed out.

5. Reinstall the A11 assembly and reconnect the coaxial cables.
Agilent Technologies 83640B — 08360125.pdf diagram (page 349)
Fig. 117 · 08360125.pdf

Figure 3-38. FM Input Impedance Jumper Locations

3-90 Adjustments · FM Input Impedance

Page 350 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 350 · click to enlarge
Page 350

B. A8 Jumper

1. On the syn thesizer, make sure the line p ow er switch is set to standb y .
2. Lift up the A8 mo dulation generator assembly far enough to access the jump er sho wn in

Figure 3-39 .

3. The FM input imp edance is factory-set to the 50 ohm p osition. T o c hange it to 600 ohms,

mov e the jump er to the p osition marked \600". The jump er can b e returned to the

p osition marked \50" to c hange the FM input imp edance to 50 ohms.

4. Reinstall the A8 assembly .
Agilent Technologies 83640B — 08360125.pdf diagram (page 350)
Fig. 118 · 08360125.pdf

Figure 3-39. A8 Jumper Locations

Related P erformance T ests

Maxim um FM Deviation

In Case of Difficulty

If the jump er is missing, see \Replaceable P arts".

Show 4 more labels / short text on this page
FM Accuracy · FM Bandwidth · Agilent 8360 Adjustments · FM Input Impedance
Page 351 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 351 · click to enlarge
Page 351
21. Modulation Generator Flatness (Option 002)

The in ternal mo dulation generator is set up for FM op eration. A 100 kHz rate, measured at

the AM/FM OUTPUT connector, is used as the reference amplitude. A calibration constant

is adjusted so the AM/FM output signal is the same amplitude as for a 1 MHz FM rate.

1. Connect the equipmen t as sho wn in Figure 3-40 . Preset all instrumen ts and let them w arm

up for at least one hour.

Agilent Technologies 83640B — 08360125.pdf diagram (page 351)
Fig. 119 · 08360125.pdf

Figure 3-40. Modulation Generator Flatness Adjustment Setup

Digitizing Oscilloscop e HP/Agilent 54111D

2. On the syn thesizer, press:

MOD

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FM On/Off Int (asterisk on)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Internal FM Rate

Internal FM Dev

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRIOR

3. On the oscilloscop e, set:

Time/Division 2 s

Trig Src Chan 1

Show 29 more labels / short text on this page
(8360 B-Series Only) · Description and Procedure · Equipment · · · FM Menu · 100 · kHz · · 8 · MHz · · · · Monitor Menu · ModOut On/Off FM · Channel 1 · Display On · Volts/Division 210 mV · Input Coupling dc · Input Impedance 50 · Timebase · Delay Reference Center · Sweep Auto · Trigger · Trigger Mode Edge · Slope Pos · 3-92 Adjustments · Modulation Generator Flatness
Page 352 of the Agilent Technologies 83640B user manual — SERVICE
Page 352 · click to enlarge
Page 352

Display Time 0.2 s

4. Adjust the oscilloscop e v ertical sensitivit y con trols to almost ll the graticule display with

the sine w av e. Do not let the sinew av e b e clipp ed. (Use the front panel k eypad b ecause the

RPG c hanges the sensitivity in steps to o large for this purp ose.)

5. Use the measure feature of the oscilloscop e to measure the amplitude of the sinew av e. (If a

Signal clipped message is display ed, rep eat the v ertical sensitivit y adjustmen t un til the

message is no longer display ed when the signal is measured.)

Record the amplitude of the sinew av e V p-p.

6. On the oscilloscop e, c hange the timebase to 200 ns/div.
7. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Internal FM Rate

SERVICE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ENTER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Modify Cal and, using the rotary knob, c hange the calibration constant un til the

sinew av e display ed on the oscilloscop e is the same amplitude (V p-p) as the recorded v alue.

(The oscilloscop e do es not automatically up date the measuremen t; con tinue to press the

measure function as y ou c hange the calibration constant.)

9. If this is the last calibration constant y ou will b e adjusting, see \Calibration Constants" to

store the calibration constants as protected data

(in EEPR OM).

Related P erformance T ests

In Case of Difficulty

Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

Show 20 more labels / short text on this page
Display · Display Mode Repetitive · Averaging Off · · · 1 · · FM Menu · MHz · · · Adjust Menu · Calib Menu · Select Cal · 274 · · Press · FM Accuracy · Agilent 8360 Adjustments · Modulation Generator Flatness
Page 353 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 353 · click to enlarge
Page 353
22. External Trigger Edge Selection

No test equipmen t is required for this pro cedure.

This pro cedure sets the external trigger to trigger on either the leading or the trailing edge

of a TTL signal. When the jump er (shown in Figure 3-41) is in the POS p osition a sweep is

triggered on the leading edge of the pulse (when the signal c hanges from 0 to 5 v olts). When

the jump er is in the NEG p osition a sweep is triggered on the trailing edge of the pulse (when

the signal c hanges from 5 to 0 v olts).

Agilent Technologies 83640B — 08360125.pdf diagram (page 353)
Fig. 120 · 08360125.pdf

Figure 3-41. External Trigger Edge Selection Location

Triggering on the Leading Edge

1. T urn the syn thesizer p ow er o and lift up the A15 assembly far enough to access the

jump er sho wn in Figure 3-41 . In some instrumen ts the jump er p ositions will b e rotated to

. In these instrumen ts the jump er v ertically connects the tw o left p oints in the

POS p osition and the tw o right p oints in the NEG p osition. (POS and NEG p ositions are

lab eled on all assemblies.)

2. The jump er is factory-set to the POS p osition (triggers on the leading edge). If the jump er

has b een mov ed to the NEG p osition mov e bac k it to the POS p osition. The syn thesizer

no w triggers a sweep on the leading edge

(a 0 to 5 v olt c hange).

3. Reinstall the A15 assembly and turn syn thesizer p ow er on.

Triggering on the Trailing Edge

1. T urn the syn thesizer p ow er o and lift up the A15 assembly far enough to access the

jump er sho wn in Figure 3-41 . In some instrumen ts the jump er p ositions will b e rotated to

. In these instrumen ts the jump er v ertically connects the tw o left p oints in the

POS p osition and the tw o right p oints in the NEG p osition. (POS and NEG p ositions are

lab eled on all assemblies.)

2. The jump er is factory-set to the POS p osition (triggers on the leading edge). T o c hange

it mov e the jump er to the NEG p osition. The syn thesizer no w triggers a sweep on the

trailing edge (a 5 to 0 v olt c hange).

3. Reinstall the A15 assembly and turn syn thesizer p ow er on.

External Trigger Edge Selection

Show 6 more labels / short text on this page
Description and Procedure · the left 90 · · the left 90 · · 3-94 Adjustments
Page 354 of the Agilent Technologies 83640B user manual — SW1 SW2 SW3 SW4 SW5
Page 354 · click to enlarge
Page 354
23. V olts/GHz

No test equipmen t is required for this pro cedure.

A v e-section switch on the A12 m ultiplier/ lter driver assembly sets the v olts/GHz scale

for the rear panel V/GHz output connector. A t the factory , syn thesizers with a maximum

frequency of 30 GHz are set to 0.5 V/GHz and higher frequency instrumen ts are set to 0.25

V/GHz. The following pro cedure describ es ho w to set the switch to scale the output.

1. Set A12S1 (shown in Figure 3-42) for the required v olts/GHz scale according to the

follo wing table:

No damage results from setting the v olts/GHz to a scale that is b ey ond the

frequency limits given in the table ho w ev er the v olts/GHz will b e inaccurate

b ey ond the frequency limits.

SW1 SW2 SW3 SW4 SW5

1 V/GHz 20 GHz closed X

0.5 V/GHz 40 GHz op en op en closed closed X

0.25 V/GHz 50 GHz op en op en op en op en X

Agilent Technologies 83640B — SW1 SW2 SW3 SW4 SW5 diagram (page 354)
Fig. 121

Figure 3-42. A12S1 Location

Related P erformance T ests

Show 9 more labels / short text on this page
Description and Procedure · V olts/GHz · Maximum · Switc h Settings · Scale · F requency · None · Agilent 8360 Adjustments · V olts/GHz
Page 355 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 355 · click to enlarge
Page 355

In Case of Difficulty

Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

3-96 Adjustments · V olts/GHz

Page 356 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 356 · click to enlarge
Page 356
24. A9 Pulse Board Jumper (W1)

No test is required

When replacing the A9 Pulse Board, the E1-E2, E3-E4 jump er needs to b e prop erly

con gured. There is one con guration for all 8360A/B mo dels. There are 2 p ossible

con gurations for 8360L mo dels. The L mo del con guration is dep enden t on the instrumen t

pre x.

On the A9 assembly , set J1 to the con guration for the sp eci c mo del as de ned in the

follo wing table:

Mo del Pre x Jump er

83623L 3722A and ab ov e E3-E4

83630L 3722A and ab ov e E3-E4

83640L 3722A and ab ov e E3-E4

83650L 3722A and ab ov e E3-E4

all other mo dels and pre x E1-E2

A9 Pulse Board Jumper (W1)

Description and Procedure · Agilent 8360 Adjustments

Page 357 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 357 · click to enlarge
Page 357
25. Selftest P atches

No test is required

After completing the A9 Pulse Board jump er selection, v erify that the following test patc hes

are set for L mo dels with the follo wing pre x and ab ov e. No patc hes are required for L mo dels

b elow the pre x listed

Mo del Pre x

83623L 3722A and ab ov e

83630L 3722A and ab ov e

83640L 3722A and ab ov e

83650L 3722A and ab ov e

107 2 3740 1740

107 5 150 7500

108 1 3740 1740

108 2 150 7500

108 3 6000 3000

108 4 3740 1740

T o Access Selftest P atc h Men u

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

En ter password if requested(factory password is 8360) and then press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Show 21 more labels / short text on this page
(8360 L-Series Only) · Description and Procedure · T est · Datum · Minimum · Maximum · Numb er · Numb er · V alue · V alue · 283 3 skip · Press · · more 1/3 · Adjust Menu · Service · Adjust Menu · more 1/2 · TestPatch Menu · 3-98 Adjustments · Selftest P atches
Page 358 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 358 · click to enlarge
Page 358

T o Add a T est P atc h

Press:

NNNNNNNNNNNNNNNNN

Specs and the following messages will app ear on the display:

TEST NUMBER TO PATCH:0 En ter the test n um b er (107), then press

TEST DATUM POINT TO PATCH:0 En ter the datum n um b er (2), then press

MIN SPECIFICATION:0 En ter minimum v alue ( 3740), then press

MAX SPECIFICATION:0 En ter maxim um v alue ( 1740), then press

T o Skip a Selftest

NNNNNNNNNNNNNNNNNNNNNNNNNN

Press:

DataSkip the follo wing message will b e display ed:

TEST NUMBER TO PATCH:0 En ter the test n um b er (108), then press

TEST DATUM POINT TO PATCH:0 En ter the datum n um b er (3), then press

T o Delete a T est P atc h

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Scroll to the test patc h to delete and press

Delete Patch .

Show 14 more labels / short text on this page
HZ sec ENTER · · HZ sec ENTER · · HZ sec ENTER · · · HZ sec ENTER · · HZ sec ENTER · · HZ sec ENTER · Agilent 8360 Adjustments · Selftest P atches
Page 359 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 359 · click to enlarge
Page 359
Page 359 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 360 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 360 · click to enlarge
Page 360

This section con tains the follo wing information on calibration constants and ho w to use them:

Chec ksum V eri cation

Calibration Constant P assword

Changing W orking Data Calibration Constants

Sa ving W orking Data Calibration Constants

Loading Protected Data Calibration Constants

Loading Default Data Calibration Constants

Calibration constants are data whic h describ e y our individual instrumen t. Calibration

constants con tain serial n um b er pre x, hardw are con guration, and b oard revision

information. Calibration constants also con tain calibration information sp eci c to the

op eration of eac h instrumen t that is used to make sure that the instrumen t meets

sp eci cations.

Changing calibration constants can cause y our instrumen t to not meet

sp eci cations.

Agilent 8360 Calibration Constants 4-1

Show 6 more labels / short text on this page
Calibration Constants · Introduction · De nition · Memory Areas · Calibration Constant Descriptions · Definition
Page 361 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 361 · click to enlarge
Page 361

The syn thesizer has three memory areas reserv ed for calibration constants:

W orking data is the set of calibration constants accessed during normal op eration and

con tains the calibration information required for optimum instrumen t p erformance. W orking

data is stored in RAM, and is maintained b y a 1 farad capacitor.

Protected data resides in EEPR OM. This calibration data is essen tially the same as w orking

data, but is not dep enden t on the capacitor. If the syn thesizer's c hec ksum test fails, or if the

capacitor b ecomes disc harged and w orking data is lost, the syn thesizer copies protected data

v alues in to w orking data RAM.

If the calibration constants are defaulted, the syn thesizer will need to b e

completely adjusted b efore it will meet its sp eci cation.

Default data resides in UVEPR OM. This data di ers from w orking and protected data in

that it cannot b e c hanged. This data represen ts a typical syn thesizer, and is not optimized

for y our instrumen t. The default calibration constants are a starting p oint for calibration.

The syn thesizer will probably not meet sp eci cations. The syn thesizer uses default data if

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

a problem exists in b oth w orking and protected data or if y ou select

Default Cal in the

calibration utility men u.

A t instrumen t p ow er on, the calibration constants stored in RAM are added and compared

with the c hec ksum v alue. If the tw o v alues match, the curren t calibration constants are

maintained in RAM.

If the RAM c hec ksum do es not v erify , the calibration constants are loaded in to RAM from

EEPR OM. These calibration constants are added and compared with their c hec ksum v alue. If

the tw o v alues match, these calibration constants (from EEPR OM) are maintained in RAM.

If the RAM c hec ksum still do es not v erify , the default calibration constants are loaded

in to RAM from UVEPR OM. A message is display ed on the syn thesizer indicating that the

default calibration constants are in use. The syn thesizer will probably not meet p erformance

sp eci cations. All the p erformance tests should b e run.

4-2 Calibration Constants Agilent 8360

Show 5 more labels / short text on this page
Memory Areas · W orking Data · Protected Data · Default Data · Checksum V erification
Page 362 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 362 · click to enlarge
Page 362

Calibration Constant P assw ord

The syn thesizer is shipp ed with a factory-set password. A password disables access to the

adjustmen t men u unless the password is en tered (see \En tering a P assword"). Calibration

constants c annot b e man ually altered without accessing the adjustmen t men u.

The follo wing is the factory-set password: 8360 Y ou can set a new password. See \Setting a

P assword".

Y ou can eliminate the password. See \Disabling a P assw ord".

If y ou ha v e forgotten the password, or if y ou require access to the calibration constant

adjustmen ts for calibration purp oses and do not kno w the password, see \Bypassing the

P assword".

Entering a P assw ord

An asterisk on the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Disable Adjust softkey in the service men u indicates that a password is

set. In order to access the adjustmen t men u y ou m ust en ter that password:

1. On the syn thesizer, set:

SERVICE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Adjust Menu .

The follo wing message is display ed: ENTER PASSWORD: 0

2. En ter the correct password using the n umeric k eypad and press

.

ENTER

The asterisk on the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Disable Adjust softkey turns o .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Adjust Menu again to access the adjustmen t men u.

En tering the password allo ws p ermanen t access to the adjustmen t men u, ev en if line p ow er is

cycled. T o disable access again, set another password.

Disabling a P assw ord

If y ou wish to eliminate a password (either the factory-set password or one y ou set y ourself ),

p erform the follo wing:

1. On the syn thesizer, set:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Disable Adjust .

SERVICE

The follo wing message is display ed: ENTER PASSWORD: 0

2. En ter the curren t password using the n umeric k eypad and press

ENTER

.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

The asterisk on the

Disable Adjust softkey turns o .

Disabling the password allo ws p ermanen t access to the adjustmen t men u, ev en if line p ow er is

cycled.

Agilent 8360 Calibration Constants 4-3

Show 5 more labels / short text on this page
· · Select · ·
Page 363 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 363 · click to enlarge
Page 363

Setting a P assw ord

If a password is already set on the syn thesizer and y ou wish to c hange it, rst follo w the

\Disabling a P assword" pro cedure. This eliminates the curren t password. Then con tinue with

this pro cedure.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

If no password is set on the syn thesizer (there is no asterisk on the

Disable Adjust softk ey),

and y ou wish to set a password, p erform the following:

1. On the syn thesizer, set:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Disable Adjust .

SERVICE

The following message is display ed: SET PASSWORD: 0

2. En ter the desired password using the n umeric k eypad. Up to 14 n umeric digits are allo w ed.

T erminate the en try b y pressing:

ENTER

.

3. A w arning message is display ed informing y ou that the adjustmen t men us will no longer b e

accessible. T o v erify that y ou do wish to set the password, select

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Yes, Confirm .

The asterisk on the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Disable Adjust softkey turns on, indicating the password is set.

4-4 Calibration Constants Agilent 8360

·

Page 364 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 364 · click to enlarge
Page 364

Bypassing the P assw ord

If y ou require access to the adjustmen t men u for calibration purp oses, a password is set, and

y ou do not kno w the password, p erform the follo wing:

1. T urn o line p ow er to the syn thesizer.
2. Op en A15S1 switch 5 (set the switch to the " 1 " p osition).
3. T urn on line p ow er to the syn thesizer and press

SERVICE

.

The asterisk on the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Disable Adjust softkey turns o enabling access to the adjustmen t

men u.

When access to the adjustmen t men u is no longer desired, enable the previously set password.

P erform the follo wing:

4. T urn o line p ow er to the syn thesizer.
5. Close A15S1 switch 5 (set the switch to the " 0 " p osition).
6. T urn on line p ow er to the syn thesizer and press

.

SERVICE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

The asterisk on the

Disable Adjust softkey turns on indicating that y ou cannot access

the adjustmen t men u.

If y ou ha v e forgotten the password and wish to set a new one, p erform steps 1 through 3 and

then steps 7 through 10.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

On the syn thesizer, select

Disable Adjust .

The follo wing message is display ed: SET PASSWORD: 0

8. En ter the desired password using the n umeric k eypad. Up to 14 n umeric digits are

allow ed. T erminate the en try b y pressing:

.

ENTER

9. A w arning message is display ed informing y ou that the adjustmen t men us will no longer

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

b e accessible. T o v erify that y ou do wish to set the password, select

Yes, Confirm .

10. Another w arning message is display ed informing y ou that the A15S1 switch 5 is set op en.

Complete steps 4 through 6 to close the switch and enable the password.

Agilent 8360 Calibration Constants 4-5

Page 365 of the Agilent Technologies 83640B user manual — PRESET
Page 365 · click to enlarge
Page 365

Changing W orking Data Calibration Constants

If y ou need to mo dify the w orking data calibration constants, the follo wing pro cedure accesses

the calibration constants and lets y ou c hange them.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

On the syn thesizer, set

SERVICE

Calib Menu .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Select Cal .

En ter the n um b er of the calibration constant y ou wish to c hange using the up/do wn

arrow k eys, the rotary knob, or the n umeric k eypad. T erminate n umeric k eypad en tries b y

.

ENTER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Modify Cal .

Change the v alue of the calibration constant using the right/left arrow k eys, the rotary

knob, or the n umeric k eypad. T erminate n umeric k eypad en tries b y pressing

ENTER

.

4. Exit the calibration men u using the

PRIOR

k ey or b y pressing

PRESET

.

Sa ving W orking Data Calibration Constants

In some adjustmen t pro cedures y ou will c hange w orking data calibration constants. The

follo wing pro cedure stores the calibration constants as protected data (in EEPR OM).

1. On the syn thesizer, set:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Cal Util Menu .

SERVICE

NNNNNNNNNNNNNNNNNNNNNNNNNN

Save Cal . A w arning is display ed informing y ou that c hanging the calibration

constants may drastically a ect instrumen t p erformance. T o v erify that y ou do wish to

NNNNNNNNNNN

sav e the new calibration constants, select

yes .

The w orking data calibration constants are no w stored as protected data

(in EEPR OM).

3. Exit the calibration utilit y men u using the

PRIOR

k ey or pressing

PRESET

.

4-6 Calibration Constants Agilent 8360

Show 10 more labels / short text on this page
· Adjust Menu · Select · pressing · Select · · · Adjust Menu · Calib Menu · Select
Page 366 of the Agilent Technologies 83640B user manual — PRESET
Page 366 · click to enlarge
Page 366

Loading Protected Data Calibration Constants

If y our w orking data calibration constants ha v e b een altered or deleted, the follo wing

pro cedure loads the protected calibration constants from EEPR OM in to w orking data

memory .

On the syn thesizer, set

SERVICE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Cal Util Menu .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Recall Cal .

A w arning is display ed informing y ou that c hanging the calibration constants may

drastically a ect instrumen t p erformance. T o v erify that y ou do wish to recall the

NNNNNNNNNNN

protected data calibration constants, select

yes .

The protected data calibration constants are no w loaded in to w orking data.

3. Exit the calibration utilit y men u using the

k ey or pressing

.

PRIOR

PRESET

Loading Default Data Calibration Constants

If y ou w an t to use the generic syn thesizer calibration constants as w orking data, the following

pro cedure loads the default calibration constants from UVEPR OM in to w orking data memory .

1. On the syn thesizer, set:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SERVICE

Cal Util Menu .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Default Cal .

A w arning is display ed informing y ou that c hanging the calibration constants may

drastically a ect instrumen t p erformance. T o v erify that y ou do wish to recall the default

NNNNNNNNNNN

yes .

The default data calibration constants are no w loaded in to w orking data.

3. Exit the calibration utilit y men u using the

PRIOR

k ey or pressing

PRESET

.

Agilent 8360 Calibration Constants 4-7

Show 10 more labels / short text on this page
· Adjust Menu · Calib Menu · Select · · · Adjust Menu · Calib Menu · Select · calibration constants, select
Page 367 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 367 · click to enlarge
Page 367

T able 4-9 provides the following information for eac h calibration constant:

The calibration constant n um b er in ascending order.

The calibration constant name. An asterisk indicates that this calibration constant is

adjusted b y an automated adjustmen t.

The range of realistic adjustmen t (adjustment outside of the range may b e p ossible but

will probably cause the syn thesizer to b e sev erely misadjusted) or the default v alue (some

calibration constants are set to a default v alue and are not adjustable).

The related adjustmen t pro cedure, man ual, automated, or front panel, that adjusts that

calibration constant.

Blank calibration constants are reserv ed for future use.

4-8 Calibration Constants Agilent 8360

Calibration Constant Descriptions

Page 368 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 368 · click to enlarge
Page 368

T able 4-9. Calibration Constant Descriptions

Num b er Description Range or

1 Identi er Resp onse 0 | HP/Agilen t 83620A

1 | HP/Agilen t 83621A

2 | HP/Agilen t 83622A

3 | HP/Agilen t 83623A

4 | HP/Agilen t 83624A

5 | HP/Agilen t 83640A

6 | HP/Agilen t 83642A

7 | HP/Agilen t 8340B

8 | HP/Agilen t 83631A

9 | HP/Agilen t 8341B

10 | HP/Agilen t 83650A

11 | HP/Agilen t 83651A

12 | HP/Agilen t 83652A

13 | Agilent 83630A

14 | HP/Agilen t 83632A

15 | Agilent 83620B

16 | Agilent 83621B

17 | Agilent 83622B

18 | Agilent 83623B

19 | Agilent 83624B

20 | Agilent 83640B

21 | Agilent 83631B

22 | Agilent 83650B

23 | Agilent 83651B

24 | Agilent 83630B

25 | Agilent 83623L

26 | Agilent 83630L

27 | Agilent 83640L

28 | Agilent 83650L

2 Dwell After RF On 2

3 User Con guration 0

4 Lo ck out F eatures 0

5 Display Field Lo ck 0

6 Retrace Dwell Time 0

16 Pro duction Info 1 0

17 Pro duction Info 2 0

18 Pro duction Info 3 Instrument sp eci c

Agilent 8360 Calibration Constants 4-9

Related · Default · Adjustment

Page 369 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 369 · click to enlarge
Page 369

T able 4-9. Calibration Constant Descriptions (continued)

Num b er Description Range or

19 Pro duction Info 4 Instrument sp eci c

20 Pro duction Info 5 Instrument sp eci c

21 Lo ck Dwell Time 2

22 Max Ph Lk W ait 100

23 YO Slew Rate 500

24 YO Kick Max 6000

25 YO Settle Rate 20

36 YO Dly T erm A Bnd 0 400 to 800 YO Delay Adjustment

37 YO Dly T erm A Bnd 1 400 to 800 YO Delay Adjustment

38 YO Dly T erm A Bnd 2 400 to 800 YO Delay Adjustment

39 YO Dly T erm A Bnd 3 400 to 800 YO Delay Adjustment

40 YO Dly T erm A Bnd 4 400 to 800 YO Delay Adjustment

41 YO Dly T erm A Bnd 5 400 to 800 YO Delay Adjustment

42 YO Dly T erm A Bnd 6 400 to 800 YO Delay Adjustment

43 YO Dly T erm A Bnd 7 400 to 800 YO Delay Adjustment

50 YO Dly T erm B Bnd 0 300 to 700 YO Delay Adjustment

51 YO Dly T erm B Bnd 1 300 to 700 YO Delay Adjustment

52 YO Dly T erm B Bnd 2 300 to 700 YO Delay Adjustment

4-10 Calibration Constants Agilent 8360

Related · Default · Adjustment

Page 370 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 370 · click to enlarge
Page 370

T able 4-9. Calibration Constant Descriptions (continued)

Num b er Description Range or

53 YO Dly T erm B Bnd 3 300 to 700 YO Delay Adjustment

54 YO Dly T erm B Bnd 4 300 to 700 YO Delay Adjustment

55 YO Dly T erm B Bnd 5 300 to 700 YO Delay Adjustment

56 YO Dly T erm B Bnd 6 300 to 700 YO Delay Adjustment

57 YO Dly T erm B Bnd 7 300 to 700 YO Delay Adjustment

72 *CW O set 500 to +500 YO Driver Gain And Linearit y

73 YO O V er; Band 0 0

74 YO O V er; Band 1 0

75 YO O V er; Band 2 0

76 YO O V er; Band 3 0

77 YO O V er; Band 4 0

78 YO O V er; Band 5 0

79 YO O V er; Band 6 0

80 YO O V er; Band 7 0

Agilent 8360 Calibration Constants 4-11

Related · Default · Adjustment

Page 371 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 371 · click to enlarge
Page 371

T able 4-9. Calibration Constant Descriptions (continued)

Num b er Description Range or

87 YO Swp V er; Band 0 0

88 YO Swp V er; Band 1 0

89 YO Swp V er; Band 2 0

90 YO Swp V er; Band 3 0

91 YO Swp V er; Band 4 0

92 YO Swp V er; Band 5 0

93 YO Swp V er; Band 6 0

94 YO Swp V er; Band 7 0

101 YTM T emp Co ecient 0

105 SRD Bias A; Band 1 255 Amp/Multiplier Adjustments

106 SRD Bias A; Band 2 30 to 230 Amp/Multiplier Adjustments

107 SRD Bias A; Band 3 30 to 230 Amp/Multiplier Adjustments

108 SRD Bias A; Band 4 30 to 230 Amp/Multiplier Adjustments

109 SRD Bias A; Band 5 30 to 230 Amp/Multiplier Adjustments

110 SRD Bias A; Band 6 30 to 230 Amp/Multiplier Adjustments

111 SRD Bias A; Band 7 30 to 230 Amp/Multiplier Adjustments

4-12 Calibration Constants Agilent 8360

Related · Default · Adjustment

Page 372 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 372 · click to enlarge
Page 372

T able 4-9. Calibration Constant Descriptions (continued)

Num b er Description Range or

118 SRD Bias B; Band 1 255 Amp/Multiplier Adjustments

119 SRD Bias B; Band 2 30 to 230 Amp/Multiplier Adjustments

120 SRD Bias B; Band 3 30 to 230 Amp/Multiplier Adjustments

121 SRD Bias B; Band 4 30 to 230 Amp/Multiplier Adjustments

122 SRD Bias B; Band 5 30 to 230 Amp/Multiplier Adjustments

123 SRD Bias B; Band 6 30 to 230 Amp/Multiplier Adjustments

124 SRD Bias B; Band 7 30 to 230 Amp/Multiplier Adjustments

131 Squegg Clamp 1A 15 to 200 Ampli er/Filter Adjustments

132 Squegg Clamp 2A 255 Ampli er/Filter Adjustments

133 Squegg Clamp 3A 255 Ampli er/Filter Adjustments

134 Squegg Clamp 4A 255 Ampli er/Filter Adjustments

135 Squegg Clamp 5A 255 Ampli er/Filter Adjustments

136 Squegg Clamp 6A 255 Ampli er/Filter Adjustments

137 Squegg Clamp 7A 255 Ampli er/Filter Adjustments

144 YTM Dly T erm A Hrm 1 600 to 2000 Ampli er/Filter Adjustments

145 YTM Dly T erm A Hrm 2 600 to 2500 Ampli er/Filter Adjustments

146 YTM Dly T erm A Hrm 3 600 to 3000 Ampli er/Filter Adjustments

147 YTM Dly T erm A Hrm 4 600 to 2500 Ampli er/Filter Adjustments

148 YTM Dly T erm A Hrm 5 600 to 3000 Ampli er/Filter Adjustments

149 YTM Dly T erm A Hrm 6 600 to 3000 Ampli er/Filter Adjustments

150 YTM Dly T erm A Hrm 7 600 to 3000 Ampli er/Filter Adjustments

Agilent 8360 Calibration Constants 4-13

Related · Default · Adjustment

Page 373 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 373 · click to enlarge
Page 373

T able 4-9. Calibration Constant Descriptions (continued)

Num b er Description Range or

157 YTM Dly T erm B Hrm 1 200 to 500 Ampli er/Filter Adjustments

158 YTM Dly T erm B Hrm 2 200 to 500 Ampli er/Filter Adjustments

159 YTM Dly T erm B Hrm 3 200 to 500 Ampli er/Filter Adjustments

160 YTM Dly T erm B Hrm 4 200 to 500 Ampli er/Filter Adjustments

161 YTM Dly T erm B Hrm 5 200 to 500 Ampli er/Filter Adjustments

162 YTM Dly T erm B Hrm 6 200 to 500 Ampli er/Filter Adjustments

163 YTM Dly T erm B Hrm 7 200 to 500 Ampli er/Filter Adjustments

170 YTM Bx Dly T erm A 1 600 to 2000 Amp/Multiplier Adjustments

171 YTM Bx Dly T erm A 2 600 to 2500 Amp/Multiplier Adjustments

172 YTM Bx Dly T erm A 3 600 to 3000 Amp/Multiplier Adjustments

173 YTM Bx Dly T erm A 4 600 to 2500 Amp/Multiplier Adjustments

174 YTM Bx Dly T erm A 5 600 to 3000 Amp/Multiplier Adjustments

175 YTM Bx Dly T erm A 6 600 to 3000 Amp/Multiplier Adjustments

176 YTM Bx Dly T erm A 7 600 to 3000 Amp/Multiplier Adjustments

4-14 Calibration Constants Agilent 8360

Related · Default · Adjustment

Page 374 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 374 · click to enlarge
Page 374

T able 4-9. Calibration Constant Descriptions (continued)

Num b er Description Range or

183 YTM Bx Dly T erm B 1 200 to 500 Amp/Multiplier Adjustments

184 YTM Bx Dly T erm B 2 200 to 500 Amp/Multiplier Adjustments

185 YTM Bx Dly T erm B 3 200 to 500 Amp/Multiplier Adjustments

186 YTM Bx Dly T erm B 4 200 to 500 Amp/Multiplier Adjustments

187 YTM Bx Dly T erm B 5 200 to 500 Amp/Multiplier Adjustments

188 YTM Bx Dly T erm B 6 200 to 500 Amp/Multiplier Adjustments

189 YTM Bx Dly T erm B 7 200 to 500 Amp/Multiplier Adjustments

195 YTM Kick Threshold 110 Amp/Multiplier Adjustments

196 YTM CW Kick Max 600 Amp/Multiplier Adjustments

197 YTM Mono Band Kick 2000 Amp/Multiplier Adjustments

198 YTM Stereo Band Kick 2000 Amp/Multiplier Adjustments

199 YTM Slew Rate 400 Amp/Multiplier Adjustments

200 YTM Slew Max 50

201 YTM Slew Min 0

202 YTM Neg Kick W ait 2

203 YTM Fwd Kick Pct 50

204 YTM Fwd Kick W ait 30

205 YTM Rise; Band 1 1 to 20 Ampli er/Filter Adjustments

206 YTM Rise; Band 2 A1 20 to 25 Ampli er/Filter Adjustments

207 YTM Rise; Band 2 B1 1 to 4 Ampli er/Filter Adjustments

208 YTM Rise; Band 2 A2 10 to 20 Ampli er/Filter Adjustments

209 YTM Rise; Band 3 A1 5 to 255 Ampli er/Filter Adjustments

210 YTM Rise; Band 3 B1 0 to 50 Ampli er/Filter Adjustments

211 YTM Rise; Band 3 A2 5 to 255 Ampli er/Filter Adjustments

212 YTM Rise; Band 3 B2 0 to 50 Ampli er/Filter Adjustments

213 YTM Rise; Band 3 A3 1 to 20 Ampli er/Filter Adjustments

Agilent 8360 Calibration Constants 4-15

Related · Default · Adjustment

Page 375 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 375 · click to enlarge
Page 375

T able 4-9. Calibration Constant Descriptions (continued)

Num b er Description Range or

214 YTM Rise; Band 4 1 to 50 Ampli er/Filter Adjustments

215 YTM Rise; Band 5 1 to 50 Ampli er/Filter Adjustments

216 YTM Rise; Band 6 1 to 50 Ampli er/Filter Adjustments

217 YTM Rise; Band 7 1 to 50 Ampli er/Filter Adjustments

225 *YTM Gain; Band 1 1500 to 2500 Amp/Multiplier Adjustments;

Auto T rac k

226 *YTM Gain; Band 2 1000 to 2500 Amp/Multiplier Adjustments;

Auto T rac k

227 *YTM Gain; Band 3 500 to 2500 Amp/Multiplier Adjustments;

Auto T rac k

228 *YTM Gain; Band 4 500 to 2500 Amp/Multiplier Adjustments;

Auto T rac k

229 *YTM Gain; Band 5 500 to 2500 Amp/Multiplier Adjustments;

Auto T rac k

230 *YTM Gain; Band 6 500 to 2500 Amp/Multiplier Adjustments;

Auto T rac k

231 *YTM Gain; Band 7 500 to 2500 Amp/Multiplier Adjustments;

Auto T rac k

238 *YTM O set; Band 1 1500 to 2500 Amp/Multiplier Adjustments;

Auto T rac k

239 *YTM O set; Band 2 1500 to 3000 Amp/Multiplier Adjustments;

Auto T rac k

4-16 Calibration Constants Agilent 8360

Related · Default · Adjustment

Page 376 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 376 · click to enlarge
Page 376

T able 4-9. Calibration Constant Descriptions (continued)

Num b er Description Range or

240 *YTM O set; Band 3 1500 to 3000 Amp/Multiplier Adjustments;

Auto T rac k

241 *YTM O set; Band 4 1500 to 3000 Amp/Multiplier Adjustments;

Auto T rac k

242 *YTM O set; Band 5 1500 to 3000 Amp/Multiplier Adjustments;

Auto T rac k

243 *YTM O set; Band 6 1500 to 3000 Amp/Multiplier Adjustments;

Auto T rac k

244 *YTM O set; Band 7 1500 to 3000 Amp/Multiplier Adjustments;

Auto T rac k

248 YTM B2 O set O set 50 to +50 Amp/Multiplier Adjustments

249 Autotrac k D A C Setting 2048

251 *L VL D A C Gain Hi Bnd 100 to +100 ALC P o wer Level

252 *L VL D A C Gain Lo Bnd 100 to +100 ALC P o wer Level

253 *L VL D A C Gain Xtal 100 to +100

254 *L VL D A C Gain 2ndOut 0

255 *L VL D A C Gain PwrMtr 100 to +100

256 *L VL D A C Gain Mo dule 100 to +100

257 *L VL D A C Gain ALC O 0

264 *L VL D A C Ofs Hi Bnd 50 to +300 ALC P o wer Level

Agilent 8360 Calibration Constants 4-17

Show 14 more labels / short text on this page
Related · Default · Adjustment · ?????????????????????? · Accuracy; · Detector Cal · ?????????????????????? · Accuracy; · Detector Cal · ?????????????????????? · Detector Cal · Accuracy n · ?????????????????????? · Detector Cal
Page 377 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 377 · click to enlarge
Page 377

T able 4-9. Calibration Constant Descriptions (continued)

Num b er Description Range or

265 *L VL D A C Ofs Lo Bnd 100 to 250 ALC P o wer Level

266 *L VL D A C Ofs Xtal 1000 to +1000

267 *L VL D A C Ofs 2ndOut 0

268 *L VL D A C Ofs PwrMtr 100 to +100

269 *L VL D A C Ofs Mo dule 100 to +100

270 *L VL D A C Ofs ALC O 0

274 In ternal FM Flat 10,000 to +10,000 Mo dulation Generator

275 FM D A C Gain T rim 1000 to +1000

276 AM D A C Gain T rim 1000 to +1000

277 AM O set 20 to +20

277 AM O set (L Mo dels) 620 to +20 AM O set (L Mo dels)

(serial pre xes 3722

283 *ALC Det Ofs; Hi 100 to 160 ALC P o wer Level

284 *ALC Det Ofs; Lo 100 to 160 ALC P o wer Level

285 *ALC Det Ofs; Xtal 100 to 160

286 *ALC Det Ofs; 2nd Out 100 to 160

292 *ALC Log Brkpt; Hi 0 to 255 ALC P o wer Level

4-18 Calibration Constants Agilent 8360

Show 18 more labels / short text on this page
Related · Default · Adjustment · ?????????????????????? · Accuracy n · Detector Cal · ?????????????????????? · Detector Cal · Flatness · ?????????????????????? · Accuracy; · Detector Cal · ?????????????????????? · Accuracy; · Detector Cal · ?????????????????????? · Accuracy; · Detector Cal
Page 378 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 378 · click to enlarge
Page 378

T able 4-9. Calibration Constant Descriptions (continued)

Num b er Description Range or

293 *ALC Log Brkpt; Lo 0 to 255 ALC P o wer Level

294 *ALC Log Brkpt; Xtal 0 to 255

295 *ALC Log Brkpt; 2nd 0 to 255

301 *ALC Mo d Gain; Band 0 100 to 200 Mo dulator O set And Gain

302 *ALC Mo d Gain; Band 1 100 to 200 Mo dulator O set And Gain

303 *ALC Mo d Gain; Band 2 70 to 200 Mo dulator O set And Gain

304 *ALC Mo d Gain; Band 3 70 to 200 Mo dulator O set And Gain

305 *ALC Mo d Gain; Band 4 100 to 200 Mo dulator O set And Gain

306 *ALC Mo d Gain; Band 5 50 to 150 Mo dulator O set And Gain

307 *ALC Mo d Gain; Band 6 50 to 150 Mo dulator O set And Gain

308 *ALC Mo d Gain; Band 7 50 to 150 Mo dulator O set And Gain

315 ALC Mo d Ofs; Band 0 50 to 150 Mo dulator O set And Gain

316 ALC Mo d Ofs; Band 1 50 to 150 Mo dulator O set And Gain

317 ALC Mo d Ofs; Band 2 150 to 220 Mo dulator O set And Gain

318 ALC Mo d Ofs; Band 3 150 to 220 Mo dulator O set And Gain

319 ALC Mo d Ofs; Band 4 100 to 200 Mo dulator O set And Gain

320 ALC Mo d Ofs; Band 5 100 to 200

321 ALC Mo d Ofs; Band 6 100 to 200

322 ALC Mo d Ofs; Band 7 100 to 200

Agilent 8360 Calibration Constants 4-19

Show 6 more labels / short text on this page
Related · Default · Adjustment · ?????????????????????? · Accuracy; · Detector Cal
Page 379 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 379 · click to enlarge
Page 379

T able 4-9. Calibration Constant Descriptions (continued)

Num b er Description Range or

329 Slo w Sym Lo w Band 100 to 160

330 Slo w Sym High Band 100 to 160

331 A ttenuator Sw P oin t 1000

335 Squegg Clamp 1B 100

336 Squegg Clamp 2B 100

337 Squegg Clamp 3B 100

338 Squegg Clamp 4B 100

339 Squegg Clamp 5B 100

340 Squegg Clamp 6B 100

341 Squegg Clamp 7B 100

351 ADC 5 V Range +OFF 5000 to +5000 ADC Adjustment

352 ADC 15 V Range +OFF 5000 to +5000 ADC Adjustment

353 ADC 5 V Range +GAIN 1950 to +1950 ADC Adjustment

354 ADC 15 V Range +GAIN 750 to +750 ADC Adjustment

4-20 Calibration Constants Agilent 8360

Related · Default · Adjustment

Page 380 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 380 · click to enlarge
Page 380

T able 4-9. Calibration Constant Descriptions (continued)

Num b er Description Range or

355 ADC 5 V Range OFF 5000 to +5000 ADC Adjustment

356 ADC 15 V Range OFF 5000 to +5000 ADC Adjustment

357 ADC 5 V Range GAIN 750 to +750 ADC Adjustment

358 ADC 15 V Range GAIN 1750 to +1950 ADC Adjustment

367 PwrMtr Gain; Lo Band 0

368 PwrMtr Gain; Hi Band 0

369 Pwr Mtr O set 0

373 YTM Dly T erm C Hrm 1 0

374 YTM Dly T erm C Hrm 2 0

375 YTM Dly T erm C Hrm 3 0

376 YTM Dly T erm C Hrm 4 0

377 YTM Dly T erm C Hrm 5 0

378 YTM Dly T erm C Hrm 6 0

379 YTM Dly T erm C Hrm 7 0

382 YTM Bx Dly T erm C 1 0

383 YTM Bx Dly T erm C 2 0

384 YTM Bx Dly T erm C 3 0

385 YTM Bx Dly T erm C 4 0

386 YTM Bx Dly T erm C 5 0

Agilent 8360 Calibration Constants 4-21

Related · Default · Adjustment

Page 381 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 381 · click to enlarge
Page 381

T able 4-9. Calibration Constant Descriptions (continued)

Num b er Description Range or

387 YTM Bx Dly T erm C 6 0

388 YTM Bx Dly T erm C 7 0

393 *Lo op Gain200.0 MHz 10 to 20 Sampler Assem bly
394 *Lo op Gain200.5 MHz 10 to 20 Sampler Assem bly
395 *Lo op Gain201.0 MHz 10 to 20 Sampler Assem bly
396 *Lo op Gain201.5 MHz 10 to 20 Sampler Assem bly
397 *Lo op Gain202.0 MHz 10 to 20 Sampler Assem bly
398 *Lo op Gain202.5 MHz 10 to 20 Sampler Assem bly
399 *Lo op Gain203.0 MHz 10 to 20 Sampler Assem bly
400 *Lo op Gain203.5 MHz 10 to 20 Sampler Assem bly
401 *Lo op Gain204.0 MHz 10 to 20 Sampler Assem bly
402 *Lo op Gain204.5 MHz 10 to 20 Sampler Assem bly
403 *Lo op Gain205.0 MHz 10 to 20 Sampler Assem bly
404 *Lo op Gain205.5 MHz 10 to 20 Sampler Assem bly
405 *Lo op Gain206.0 MHz 10 to 20 Sampler Assem bly
406 *Lo op Gain206.5 MHz 10 to 20 Sampler Assem bly
407 *Lo op Gain207.0 MHz 10 to 20 Sampler Assem bly
408 *Lo op Gain207.5 MHz 10 to 20 Sampler Assem bly
409 *Lo op Gain208.0 MHz 10 to 20 Sampler Assem bly
410 *Lo op Gain208.5 MHz 10 to 20 Sampler Assem bly
411 *Lo op Gain209.0 MHz 10 to 20 Sampler Assem bly
412 *Lo op Gain209.5 MHz 10 to 20 Sampler Assem bly
413 *Lo op Gain210.0 MHz 10 to 20 Sampler Assem bly
414 *Lo op Gain210.5 MHz 10 to 20 Sampler Assem bly
415 *Lo op Gain211.0 MHz 10 to 20 Sampler Assem bly
416 *Lo op Gain211.5 MHz 12 to 24 Sampler Assem bly
417 *Lo op Gain212.0 MHz 13 to 26 Sampler Assem bly
418 *Lo op Gain212.5 MHz 14 to 28 Sampler Assem bly

4-22 Calibration Constants Agilent 8360

Related · Default · Adjustment

Page 382 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 382 · click to enlarge
Page 382

T able 4-9. Calibration Constant Descriptions (continued)

Num b er Description Range or

419 *Lo op Gain213.0 MHz 14 to 28 Sampler Assem bly
420 *Lo op Gain213.5 MHz 15 to 30 Sampler Assem bly
421 *Lo op Gain214.0 MHz 16 to 32 Sampler Assem bly
422 *Lo op Gain214.5 MHz 17 to 34 Sampler Assem bly
423 *Lo op Gain215.0 MHz 18 to 36 Sampler Assem bly
424 *Lo op Gain215.5 MHz 18 to 36 Sampler Assem bly
425 *Lo op Gain216.0 MHz 19 to 38 Sampler Assem bly
426 *Lo op Gain216.5 MHz 20 to 40 Sampler Assem bly
427 *Lo op Gain217.0 MHz 22 to 44 Sampler Assem bly
428 *Lo op Gain217.5 MHz 23 to 46 Sampler Assem bly
429 *Lo op Gain218.0 MHz 24 to 48 Sampler Assem bly
430 *Lo op Gain218.5 MHz 28 to 56 Sampler Assem bly
431 *Lo op Gain219.0 MHz 30 to 60 Sampler Assem bly
432 *Lo op Gain219.5 MHz 32 to 64 Sampler Assem bly
433 *Lo op Gain220.0 MHz 32 to 64 Sampler Assem bly

440 Doubler Amp Mo de 0

441 Ph uo c Magic Enable 0

442 *Sweeptime Range 0 3750 to 6250 Sweep Ramp

443 *Sweeptime Range 1 19655 to 32755 Sweep Ramp

444 *Sweeptime Range 2 9360 to 15600 Sweep Ramp

449 D Amp Clamp 1 75 to 250 Ampli er/Filter Adjustments

450 D Amp Clamp 2 75 to 250 Ampli er/Filter Adjustments

Agilent 8360 Calibration Constants 4-23

Related · Default · Adjustment

Page 383 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 383 · click to enlarge
Page 383

T able 4-9. Calibration Constant Descriptions (continued)

Num b er Description Range or

451 D Amp Clamp 3 75 to 250 Ampli er/Filter Adjustments

452 D Amp Clamp 4 75 to 250 Ampli er/Filter Adjustments

453 D Amp Clamp 5 75 to 255

454 D Amp Clamp 6 75 to 255

455 D Amp Clamp 7 75 to 255

461 Serial Num b Pre x See fo otnote 1

463 Hardw are Con g 1 See fo otnote 1

464 Hardw are Con g 2 See fo otnote 1

473 P o wer Sp ec Band 0 See fo otnote 2

474 P o wer Sp ec Band 1 See fo otnote 2

475 P o wer Sp ec Band 2 See fo otnote 2

476 P o wer Sp ec Band 3 See fo otnote 2

477 P o wer Sp ec Band 4 See fo otnote 2

478 P o wer Sp ec Band 5 See fo otnote 2

479 P o wer Sp ec Band 6 See fo otnote 2

480 P o wer Sp ec Band 7 See fo otnote 2

4-24 Calibration Constants Agilent 8360

Show 4 more labels / short text on this page
Related · Default · Adjustment · 460 Debugs 0
Page 384 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 384 · click to enlarge
Page 384

T able 4-9. Calibration Constant Descriptions (continued)

Num b er Description Range or

487 Max Sweep Rate 300

488 Max Sweep Rate 8516 100

493 A1 Revision See fo otnote 3

494 A2 Revision See fo otnote 3

495 A3 Revision See fo otnote 3

496 A4 Revision See fo otnote 3

497 A5 Revision See fo otnote 3

498 A6 Revision See fo otnote 3

499 A7 Revision See fo otnote 3

500 A8 Revision See fo otnote 3

501 A9 Revision See fo otnote 3

502 A10 Revision See fo otnote 3

503 A11 Revision See fo otnote 3

504 A12 Revision See fo otnote 3

505 A13 Revision See fo otnote 3

506 A14 Revision See fo otnote 3

507 A15 Revision See fo otnote 3

508 A16 Revision See fo otnote 3

509 A17 Revision See fo otnote 3

510 A18 Revision See fo otnote 3

511 A19 Revision See fo otnote 3

512 A20 Revision See fo otnote 3

513 A21 Revision See fo otnote 3

514 A22 Revision See fo otnote 3

515 A23 Revision See fo otnote 3

Agilent 8360 Calibration Constants 4-25

Related · Default · Adjustment

Page 385 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 385 · click to enlarge
Page 385

T able 4-9. Calibration Constant Descriptions (continued)

Num b er Description Range or

525 Hib ernation Time 90

526 Bandcross Time 1 42

527 Bandcross Time 2 32

528 Bandcross Time 3 28

529 Bandcross Time 4 26

530 Bandcross Time 5 28

531 Bandcross Time 6 20

532 Bandcross Time 7 20

536 Sp ecial Option 1 0

537 Sp ecial Option 2 0

538 Sp ecial Option 3 0

539 Sp ecial Option 4 0

540 Sp ecial Option 5 0

541 Sp ecial Option 6 0

542 Sp ecial Option 7 0

543 Sp ecial Option 8 0

544 Sp ecial Option 9 0

545 Sp ecial Option 10 0

4-26 Calibration Constants Agilent 8360

Related · Default · Adjustment

Page 386 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 386 · click to enlarge
Page 386

T able 4-9. Calibration Constant Descriptions (continued)

Num b er Description Range or

554 Display FIFO Time 245

555 Display FSCC Time 1

561 YTF Dly T erm A Hrm 1 1000

562 YTF Dly T erm A Hrm 2 1000

563 YTF Dly T erm A Hrm 3 1000

564 YTF Dly T erm A Hrm 4 1000

565 YTF Dly T erm A Hrm 5 1000

566 YTF Dly T erm A Hrm 6 1000

567 YTF Dly T erm A Hrm 7 1000

571 YTF Dly T erm B Hrm 1 300

572 YTF Dly T erm B Hrm 2 300

573 YTF Dly T erm B Hrm 3 300

574 YTF Dly T erm B Hrm 4 300

575 YTF Dly T erm B Hrm 5 300

576 YTF Dly T erm B Hrm 6 300

577 YTF Dly T erm B Hrm 7 300

Agilent 8360 Calibration Constants 4-27

Related · Default · Adjustment

Page 387 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 387 · click to enlarge
Page 387

T able 4-9. Calibration Constant Descriptions (continued)

Num b er Description Range or

581 YTF Dly T erm C Hrm 1 0

582 YTF Dly T erm C Hrm 2 0

583 YTF Dly T erm C Hrm 3 0

584 YTF Dly T erm C Hrm 4 0

585 YTF Dly T erm C Hrm 5 0

586 YTF Dly T erm C Hrm 6 0

587 YTF Dly T erm C Hrm 7 0

591 YTF Bx Dly T erm A 1 1000

592 YTF Bx Dly T erm A 2 1000

593 YTF Bx Dly T erm A 3 600 to 2500

594 YTF Bx Dly T erm A 4 600 to 2500

595 YTF Bx Dly T erm A 5 1000

596 YTF Bx Dly T erm A 6 1000

597 YTF Bx Dly T erm A 7 1000

601 YTF Bx Dly T erm B 1 300 Ampli er/Filter Adjustments

602 YTF Bx Dly T erm B 2 300 Ampli er/Filter Adjustments

603 YTF Bx Dly T erm B 3 200 to 450 Ampli er/Filter Adjustments

604 YTF Bx Dly T erm B 4 200 to 450 Ampli er/Filter Adjustments

605 YTF Bx Dly T erm B 5 300 Ampli er/Filter Adjustments

606 YTF Bx Dly T erm B 6 300 Ampli er/Filter Adjustments

607 YTF Bx Dly T erm B 7 300 Ampli er/Filter Adjustments

4-28 Calibration Constants Agilent 8360

Related · Default · Adjustment

Page 388 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 388 · click to enlarge
Page 388

T able 4-9. Calibration Constant Descriptions (continued)

Num b er Description Range or

611 YTF Bx Dly T erm C 1 0

612 YTF Bx Dly T erm C 2 0

613 YTF Bx Dly T erm C 3 0

614 YTF Bx Dly T erm C 4 0

615 YTF Bx Dly T erm C 5 0

616 YTF Bx Dly T erm C 6 0

617 YTF Bx Dly T erm C 7 0

621 *YTF Gain; Band 1 1500 to 2500 Ampli er/Filter Adjustments

622 *YTF Gain; Band 2 1000 to 2500 Ampli er/Filter Adjustments

623 *YTF Gain; Band 3 500 to 2500 Ampli er/Filter Adjustments

624 *YTF Gain; Band 4 500 to 2500 Ampli er/Filter Adjustments

625 *YTF Gain; Band 5 500 to 2500 Ampli er/Filter Adjustments

626 *YTF Gain; Band 6 500 to 2500 Ampli er/Filter Adjustments

627 *YTF Gain; Band 7 500 to 2500 Ampli er/Filter Adjustments

631 *YTF O set; Band 1 1500 to 2500 Ampli er/Filter Adjustments

632 *YTF O set; Band 2 1500 to 3000 Ampli er/Filter Adjustments

633 *YTF O set; Band 3 1500 to 3000 Ampli er/Filter Adjustments

634 *YTF O set; Band 4 1500 to 3000 Ampli er/Filter Adjustments

635 *YTF O set; Band 5 1500 to 3000 Ampli er/Filter Adjustments

636 *YTF O set; Band 6 1500 to 3000 Ampli er/Filter Adjustments

637 *YTF O set; Band 7 1500 to 3000 Ampli er/Filter Adjustments

Agilent 8360 Calibration Constants 4-29

Related · Default · Adjustment

Page 389 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 389 · click to enlarge
Page 389

T able 4-9. Calibration Constant Descriptions (continued)

Num b er Description Range or

641 A24 Rev;Low Band See fo otnote 3

642 A25 Rev;Mo d Splitter See fo otnote 3

643 A26 Rev;YIG Osc See fo otnote 3

644 A27 Rev;Isolator See fo otnote 3

645 A28 Rev;SYTM See fo otnote 3

646 A29 Rev;Amp/Detector See fo otnote 3

647 A30 Rev;Coupler See fo otnote 3

648 A31 Rev;Attenuator See fo otnote 3

649 A32 Rev;Doubler See fo otnote 3

650 A33 Rev;Ampli er See fo otnote 3

651 A34 Rev;RP Assem bly See fo otnote 3

652 A35 Rev;Isolator See fo otnote 3

653 A36 PM/A42 PMSF Rev See fo otnote 3

655 A38 Rev;Dual Mo d See fo otnote 3

660 Quick Step Slew 0 120 | A13 Revision=4

5 | All others

661 Quick Step Slew 1 120 | A13 Revision=4

5 | All others

662 Quick Step Slew 2 60 | A13 Revision=4

10 | All others

663 Quick Step Slew 3 40 | A13 Revision=4

15 | All others

664 Quick Step Slew 4 30 | A13 Revision=4

17 | All others

665 Quick Step Slew 5 20 | A13 Revision=4

26 | All others

666 Quick Step Slew 6 20 | A13 Revision=4

26 | All others

667 Quick Step Slew 7 15 | A13 Revision=4

35 | All others

4-30 Calibration Constants Agilent 8360

Related · Default · Adjustment

Page 390 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 390 · click to enlarge
Page 390

T able 4-9. Calibration Constant Descriptions (continued)

Num b er Description Range or

670 RF In terface ID 0

671 A9 Pulse Board ID See fo otnote 3

672 A12 YTM Driver ID See fo otnote 3

673 A28 Amp/Mult ID See fo otnote 3

674 A29 Amp/Filt ID See fo otnote 3

675 A38 Dual Mo d ID See fo otnote 3

676 A36 PM/A42 PMSF ID See fo otnote 3

677 A32 Doubler ID See fo otnote 3

678 A10 ALC Board ID See fo otnote 3

Sp eci c to the syn thesizer's hardw are con guration. The calibration constant v alue is set at

the factory . See ***: undefined*** .

Sp eci ed maximum lev eled p ow er for the band (for example, 7 dBm equals 700, 10 dBm

equals 1000).

Sp eci c to the syn thesizer's hardw are con guration. The calibration constant ID n um b er

and revision v ary dep ending on the assembly v ersion. See T able 4-11 .

Agilent 8360 Calibration Constants 4-31

Related · Default · Adjustment

Page 391 of the Agilent Technologies 83640B user manual — HP 83620B + 001 + 004 = ?
Page 391 · click to enlarge
Page 391

T able 4-10. Hardw are Configuration Cal Constants #463 and #464

Agilent Technologies 83640B — 08360125.pdf diagram (page 391)
Fig. 122 · 08360125.pdf

83620B +19,600 128 +4 +4096 +2 { +1 +2

83622B +19,584 128 +4 +4096 +2 { +1 +2

83623B +19,608 128 +4 +4096 6 127 +1 +2

83624B +19,592 128 +4 +4096 6 { +1 +2

83630B +19,600 128 +4 +4096 +2 { +1 +2

83640B +19,632 128 +4 +4096 +2 127 +1 +2

83650B +19,888 128 +4 +4096 +2 { +1 +2

83621B +30,864 {

83631B +30,864 {

83651B +32,176 {

83623L +27,792 128 { +4096 { { +1 {

83630L +27,792 128 { +4096 { { +1 {

83640L +27,824 128 { +4096 { { +1 {

83650L +28,080 128 { +4096 { { +1 {

1 Option H10 and all other options are m utually exclusiv e.

ExampleCalibration constant #463

HP 83620B + 001 + 004 = ?

19,600 - 128 + 4096 = 23,568

4-32 Calibration Constants Agilent 8360

Page 392 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 392 · click to enlarge
Page 392

T able 4-11.

Assemblies and Corresponding Cal Constant ID Numbers and Revisions

Agilent Technologies 83640B — 08360125.pdf diagram (page 392)
Fig. 123 · 08360125.pdf

Assembly ID Numb er Revision Part Numb er Description

Agilent 8360 Calibration Constants 4-33

Page 393 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 393 · click to enlarge
Page 393

T able 4-11.

Assemblies and Corresponding Cal Constant ID Numbers and Revisions

Agilent Technologies 83640B — 08360125.pdf diagram (page 393)
Fig. 124 · 08360125.pdf

Assembly ID Numb er Revision Part Numb er Description

1 08360-60264 CPU (Battery) 2 R OM Chip Set

4-34 Calibration Constants Agilent 8360

(continued)

Page 394 of the Agilent Technologies 83640B user manual — ( 83630B/L, 83640B/L,
Page 394 · click to enlarge
Page 394

T able 4-11.

Assemblies and Corresponding Cal Constant ID Numbers and Revisions

Assembly ID Numb er Revision Part Numb er Description

A28 #673 #645

1 0 5086-7596 Switched Amp/Mult ( 83621B/31B)

2 1 5086-7622 Amp/Mult ( 83630B/L

2 2 5086-7622 Amp/Mult with Aux Out ( 83620B,

83622B, 83623B /L, 83624B,

A29 #674 #646

1 1 5086-7623 Amp/Filter

( 83630B/L, 83640B/L,

2 1 5086-7636 Amp/Filter

( 83630B/40B/50B Option 006)

3 0 5086-7616 Amp/Filter ( 83620B/22B/23B/24B)

4 1 5086-7637 Amp/Filter ( 83623L)

0 0955-0148 Directional Coupler

( 83623B/L and 83624B)

0 5086-7645 Bridge Detector

( 83620B/22B and 83630B/L)

0 0955-0125 Directional Coupler ( 83631B )

0 5086-7656 Bridge Detector

( 83640B/L and 83650B/L)

0 0955-0523 Directional Coupler ( 83651B )

0 33326-60003 Step A tten uator ( 83620B/22B/24B,

0 33326-60002 Step A tten uator ( 83640B/L)

0 33326-60005 Step A tten uator ( 83650B/L)

Agilent 8360 Calibration Constants 4-35

Show 7 more labels / short text on this page
(continued) · and 83650B /L) · 83640B/L, and 83651B) · 83650B/L, and 83651B) · A30 None #647 · A31 None #648 · 83623B/L, and 83630B/L)
Page 395 of the Agilent Technologies 83640B user manual — 83622B, 83623B/L,
Page 395 · click to enlarge
Page 395

T able 4-11.

Assemblies and Corresponding Cal Constant ID Numbers and Revisions

Assembly ID Numb er Revision Part Numb er Description

A32 #677 #649

0 1 5086-7607 F requency Doubler ( 83640B /L)

1 1 5086-7604 F requency Doubler ( 83650B /L)

1 1 5086-7513 F requency Doubler ( 83651B )

0 5086-7465 Ampli er (High P ow er) ( 83623B )

0 5086-7510 Ampli er (High P ow er) ( 83624B )

0 08360-60061 RP Floating BNC (Option 004)

0 08360-60186 RP Floating BNC (Option 002 and 004)

A36 #676 #653

A37

A38 #675 #655

1 0 5086-7602 Dual Mo dulator ( 83620B,

83622B, 83623B/L,

83624B, 83630B/L,

2 0 5086-7634 AM Mo dulator

(Option 006 and 83621B/31B/51B)

2 0 5087-7017 Mo d Doubler Amp

4-36 Calibration Constants Agilent 8360

Show 8 more labels / short text on this page
(continued) · A33 None #650 · A34 None #651 · A35 None #652 · 0 Not Used · 0 Not Used · 83640B/L, and 83650B/L) · (Agilen t 83630L/40L/50L)
Page 396 of the Agilent Technologies 83640B user manual — A42 #676 #653
Page 396 · click to enlarge
Page 396

T able 4-11.

Assemblies and Corresponding Cal Constant ID Numbers and Revisions

Assembly ID Numb er Revision Part Numb er Description

A42 #676 #653

0 0 ( 83620B/22B/24B

1 1 ( 83630B/L, 83640B/L,

0 1 5086-7628 Pulse Mo d Switched Filter

( 83620B/22B Option 006)

0 1 5086-7629 Pulse Mo d Switched Filter

( 83623B/24B Option 006)

1 1 5086-7630 Pulse Mo d Switched Filter

( 83630B Option 006)

1 1 5086-7631 Pulse Mo d Switched Filter

( 83640B/50B Option 006)

Agilent 8360 Calibration Constants 4-37

(continued) · and 83623B /L) · and 83650B /L)

Page 397 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 397 · click to enlarge
Page 397
Page 397 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 398 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 398 · click to enlarge
Page 398

This c hapter explains ho w to load and run the \Agilent 8360 Service Support Softw are"

revision A.02.00. These tests require op erator in teraction.

This softw are requires that the syn thesizer is set to the SCPI programming language. This

is set b y the rear panel switch or b y the front panel. Refer to the \Installation" c hapter of

y our syn thesizer's user's guide for further information. After using the softw are, return the

syn thesizer to its original setting.

This c hapter con tains the follo wing information on using the softw are:

Setting Up the System

Installing the Softw are

Running the Softw are

Automated P erformance T ests

Step A tten uator Flatness T est

P ow er Flatness and Accuracy T est

Step A tten uator Flatness Adjustmen t

YO Delay Adjustmen t

P ow er Flatness Adjustmen t

F ront P anel Em ulation

P ow er Sensor Con guration and Calibration F actor File

Agilent 8360 Automated T ests 5-1

Show 7 more labels / short text on this page
Automated T ests · Introduction · Automated Adjustmen ts · ADC Adjustmen t · Utilities · Calibration Constants · Self-T est
Page 399 of the Agilent Technologies 83640B user manual — CR TX HFS
Page 399 · click to enlarge
Page 399

Setting Up the System

The automated tests require an HP 9000 series 200/300 desktop computer with at least 2.25

megabytes of RAM, a disk drive, and an GPIB in terface. This program will not run with

high resolution color monitors ( > 512 x 390 pixels). Softw are is provided on tw o 3.5 inc h

disks, formatted double-sided, and are usable in double-sided disk drives only . An y required

measuremen t instrumen ts are listed in eac h test pro cedure.

The test softw are requires BASIC v ersion 5.1 and the following binaries.

T able 5-1. Required Binaries

Language Extensions Driv ers

CLOCK CR T A or CR TB (dep ends on CR T)

COMPLEX CS80 or DISC (dep ends on disk)

CR TX HFS

EDIT HPIB

ERR SRM

GRAPH

GRAPHX

IO

KBD

MA T

MS

PDEV

TRANS

XREF

Refer to the BASIC user's do cumen tation for instructions on loading the op erating system

and binaries.

5-2 Automated T ests Agilent 8360

Hardw are Requirements · Operating System Requirements

Page 400 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 400 · click to enlarge
Page 400

Installing the Softw are

The automated tests are provided on tw o double-sided disks. Previous revisions may b e

di eren t.

Mak e W orking Copies

Before doing an ything else, make a w orking cop y of the master disks!

T yp eINITIALIZE \Address of drive con taining blank opp y"

The master disk is shipp ed from the factory write-protected and cannot b e written to or

initialized in this mo de. W e recommend y ou maintain this master disk in write-protect mo de.

During execution of some tests, the program reads from and writes to the disk, use a w orking

cop y that is not write-protected when y ou run the softw are.

Installing the Program onto a Hard Disk

W e recommend that y ou run the softw are from a hard disk. It is p ossible to op erate it from a

single or dual opp y disk drive, ho w ev er sp eed will b e sacri ced.

T o install the program on to y our hard disk, create a directory that will con tain the program

and its asso ciated les. (Refer to the BASIC user's do cumen tation for instructions on creating

directories.) After y ou create the directory , run the install program b elow to load all of the

les on to y our hard disk.

1. Insert master disk #1 in to the opp y disk drive.
2. T yp e: COPY \OPV 8360:,700,0" TO \/TEST/OPV 8360:,1400,0"
3. Remo v e master disk #1 and insert master disk #2 in to the disk drive.
4. T yp e LO AD \COPY DISK:,700,0"
5. Press R UN and follow the prompts to cop y all les from the second master disk to the hard

disk. In this example, the source lo cation is :,700,0 and the destination lo cation is :,1400,0

Y our hard disk is no w set up to run the automated tests.

Operating from a Flopp y Disk

When y ou run the softw are from a opp y disk drive, b e sure to leav e the second master disk

in the disk drive while the program runs. If this is not done, the program will not run.

Agilent 8360 Automated T ests 5-3

Disk Files

Page 401 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNN
Page 401 · click to enlarge
Page 401

Running the Softw are

CA UTION

This softw are uses RAM memory v olumes for fast access of data les. These

v olumes may also b e used b y other programs and could con tain data that will

b e erased b y this program. Make sure that the computer y ou are using do es

not ha v e imp ortant data in an y memory v olumes b efore running this softw are.

The computer con taining the softw are m ust b e the only con troller on the bus. If more than

one con troller is presen t, the softw are will not run prop erly .

Loading the Softw are

1. Make sure the softw are has b een copied in to a directory (if running from a hard disk) or

copied to a w ork disk (if running from a opp y disk).

2. Set the default mass storage to the directory or opp y disk that con tains the test softw are.

Use the BASIC MSI command. See the BASIC L anguage R efer enc e for more information

on setting the default mass storage.

CAPS LOCK ON or OFF cannot b e c hanged while the program is running.

Set the k eyb oard b efore y ou load the softw are.

3. Load the test softw are.

Insert master disk #1 in y our default drive, or from y our hard-driv e directory , p erform the

follo wingT yp e:

LO AD \OPV 8360"

.

RETURN

4. Press

RUN

to start the test program. If y ou are using opp y disks, the program will

prompt y ou to install the second disk.

5. Some imp ortant messages are display ed, then the program displays a men u with the

av ailable mo del n um b ers. Using the arrow k eys, select the correct mo del to b e tested.

.

SELECT

6. The program asks for the installed options and the serial n um b er of the unit under test.

En ter the appropriate information.

7. A men u of the av ailable tests is display ed. Using the arrow k eys, select the appropriate

test. Press

SELECT

.

Refer to the follo wing pages for a detailed description and sp eci c op erating

information (if an y) for a sp eci c test.

8. F ollo w the prompts in the individual tests for connection instructions and other relev an t

test information. When the test nishes, y ou are returned to the test men u.

Either select another test to run or select the

NNNNNNNNNNNNNN

DONE softkey to exit the program.

5-4 Automated T ests Agilent 8360

Configuration Limitations · Press · Press

Page 402 of the Agilent Technologies 83640B user manual — A UTOMA TED PERF ORMANCE TESTS
Page 402 · click to enlarge
Page 402

A UTOMA TED PERF ORMANCE TESTS

1. Step A ttenuator Flatness T est

P erformance v eri cation test for Option 001

A p ow er meter is used to make relative p ow er measuremen ts to determine the actual

attenuation of eac h attenuator card. Mismatch error is minimized b y including a xed

attenuation in front of the card b eing measured for b oth parts of the relativ e measuremen t

(with/without the card). The softw are uses the v alue of eac h attenuator card to calculate

what the actual attenuation is when set to a sp eci c v alue. F or example, the attenuator v alue

for the 30 dB setting is the sum of the 10 dB and the 20 dB cards. The e ect of digital ALC

correction is accoun ted for b y the softw are.

The follo wing test equipmen t is required for this p erformance test. No substitutions are

allo w ed.

T able 5-2. T est Equipment Required

10 dB A tten uator 8493C

30 dB A tten uator 11708A

P ow er Meter 438A

P ow er Sensor (50 MHz to 26.5 GHz) 8485A

P ow er Sensor (50 MHz to 26.5 GHz) 8485D

P ow er Sensor (100 kHz to 4.2 GHz) 8482A

1 Substitute HP/Agilent 8487A/D for 40 GHz synthesizers.

Agilent 8360 Automated T ests 5-5

Show 4 more labels / short text on this page
Description · Instrumen t HP/Agilent · Mo del · Numb er
Page 403 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 403 · click to enlarge
Page 403

Connect the equipmen t as sho wn in Figure 5-1 . Preset all instrumen ts and let them w arm up

for at least one hour.

This program requires that p ow er sensor calibration factors b e stored in a le.

If the calibration factors ha v e not b een en tered previously , refer to the utility ,

\4. P ow er Sensor Con guration and Calibration F actor File".

Select the step attenuator atness test and follow the prompts on the display .

Agilent Technologies 83640B — 08360125.pdf diagram (page 403)
Fig. 125 · 08360125.pdf

Figure 5-1. Step A ttenuator Flatness T est Setup

5-6 Automated T ests Agilent 8360

Procedure

Page 404 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNN
Page 404 · click to enlarge
Page 404
2. P o w er Flatness and Accuracy T est

P erformance v eri cation test for standard instrumen t

If Option 001 is installed in y our instrumen t, p erform the \Step A tten uator

Flatness" test.

This test uses the user atness correction array to measure p ow er atness at the RF output.

The p ow er is measured ev ery 10 MHz in low band and ev ery 100 MHz in high and millimeter

bands.

The calibration factors for the p ow er sensor m ust b e added to the program b efore the test

is run. Refer to the utility , \4. P ow er Sensor Con guration and Calibration F actor File", if

necessary .

The follo wing test equipmen t is required for this test. No substitutions are allo w ed. (The test

setup is given in the softw are.)

T est Equipment Required

P ow er Meter 438A

P ow er Sensor (100 kHz to 4.2 GHz) 8482A

P ow er Sensor (50 MHz to 26.5 GHz) 8485A

P ow er Sensor (50 MHz to 50 GHz)

8487A

10 dB A tten uator (DC to 26.5 GHz)

1 F or 40 GHz synthesizers only .

2 F or high p ow er synthesizers only .

Select the p ow er atness and accuracy test and en ter y our p ow er sensor data b y p erforming

the follo wing steps.

If the p ow er sensor data w as added previously , and is still curren t, skip the

follo wing steps and run the test.

1. Select the

NNNNNNNNNNNNNN

HELP softkey to access the p ow er sensor con guration and calibration men us.

2. Select p ow er meter con guration.
3. Select the sensor to edit.
4. En ter the p ow er sensor con guration data (follo w the prompts on the display).

Agilent 8360 Automated T ests 5-7

Show 5 more labels / short text on this page
Description · Instrumen t HP/Agilent · Mo del · Numb er · Procedure
Page 405 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNN
Page 405 · click to enlarge
Page 405

Do not edit the \Sensor ID". The names set at the factory m ust remain

unc hanged for the program to run prop erly . The factory recommended

\Zero Hr" is 1.00 and the \Cal Hr" is 24.00. Use the left and right arrows to

mov e the cursor within a eld. Use the up and do wn arrows to incremen t or

decremen t the v alue.

NNNNNNNNNNNNNNNNNNNN

RETURN selects the eld.

5. When all c hanges ha v e b een made, select sav e p ow er meter con guration.
6. Next select p ow er meter calibration factors to add the p ow er sensor calibration factors to

the program.

7. Select curren t to c hange the active p ow er sensor to the one for whic h y ou w ould like to

en ter the data.

8. Select edit header, and en ter the serial n um b er of the p ow er sensor.
9. Select edit calibration factors.

5-8 Automated T ests Agilent 8360

Page 406 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 406 · click to enlarge
Page 406
10. En ter the calibration factors from y our p ow er sensor (follow the prompts on the display).
11. When all of the calibration factors ha v e b een added, select store calibration factors to

store the calibration factors for future use.

12. Exit the p ow er meter calibration.
13. Run the p ow er atness and accuracy test.

Agilent 8360 Automated T ests 5-9

Page 407 of the Agilent Technologies 83640B user manual — A UTOMA TED ADJUSTMENTS
Page 407 · click to enlarge
Page 407

A UTOMA TED ADJUSTMENTS

1. Step A ttenuator Flatness Adjustment

Calibration for Option 001

Before doing an y adjustmen t, en ter the password for access to the adjustmen t calibration

constants. Use the front panel or the front panel em ulation program provided with this

softw are.

See Chapter 4 for information on enabling the calibration constants with the password.

A p ow er meter is used to make relative p ow er measuremen ts to determine the actual

attenuation of eac h attenuator card. Mismatch error is minimized b y including a xed

attenuation in front of the card b eing measured for b oth parts of the relative measuremen t

(with/without the card). The softw are uses the v alue of eac h attenuator card to calculate

what the actual attenuation is when set to a sp eci c v alue. F or example, the attenuator v alue

for the 30 dB setting is the sum of the 10 dB and the 20 dB cards. The e ect of digital ALC

correction is accoun ted for b y the softw are.

The following test equipmen t is required for this adjustmen t. No substitutions are allo w ed.

T able 5-3. T est Equipment Required

10 dB A tten uator 8493C

30 dB A tten uator 11708A

P ow er Meter 438A

P ow er Sensor (50 MHz to 26.5 GHz) 8485A

P ow er Sensor (50 MHz to 26.5 GHz) 8485D

P ow er Sensor (100 kHz to 4.2 GHz) 8482A

1 Substitute HP/Agilent 8487A/D for 40 GHz synthesizers.

Connect the equipmen t as sho wn in Figure 5-2 . Preset all instrumen ts and let them w arm up

for at least one hour.

Select the step attenuator atness adjustmen t and follow the prompts on the display .

5-10 Automated T ests Agilent 8360

Show 5 more labels / short text on this page
Description · Instrument HP/Agilent · Mo del · Numb er · Procedure
Page 408 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 408 · click to enlarge
Page 408

Figure 5-2. Step A ttenuator Flatness Adjustment Setup

Agilent 8360 Automated T ests 5-11

Page 409 of the Agilent Technologies 83640B user manual — CA UTION
Page 409 · click to enlarge
Page 409
2. YO Delay Adjustment

This adjustmen t uses the in ternal coun ter to measure swept frequency accuracy . YO dela y

a ects the swept frequency accuracy at fast sweep times (less than 100 ms). This pro cedure

adjusts the calibration constants for the YIG oscillator to ac hiev e consistent swept frequency

accuracy ov er the en tire op erating range of the instrumen t.

CA UTION

If y our instrumen t has Option 006, b e certain that y ou iden ti ed y our option

con guration when the softw are w as initially loaded or else this pro cedure will

not run prop erly .

Select the YO dela y adjustmen t and follo w the prompts on the display .

No test equipmen t is required.

5-12 Automated T ests Agilent 8360

Description

Page 410 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 410 · click to enlarge
Page 410
3. ADC Adjustment

This adjustmen t uses an external digital v oltmeter (DVM) to measure V COMP on the YO

driver assembly and compares the reading to the ADC measuremen t. The ADC calibration

constants are adjusted so that the DVM and ADC measuremen ts are the same. The

pro cedure is rep eated for eac h v oltage range (b oth plus and minus supplies).

Select the ADC adjustmen t and follo w the prompts on the display .

The follo wing test equipmen t is required for this adjustmen t. (The test setup is given in the

softw are.)

T est Equipment Required

Instrumen t HP/Agilent Mo del Numb er

Digital V oltmeter 3456A, 3457A, or 3458A

Agilent 8360 Automated T ests 5-13

Description

Page 411 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNN
Page 411 · click to enlarge
Page 411
4. P o w er Flatness Adjustment

Calibration for standard instrumen t

If Option 001 is installed in y our instrumen t, p erform the \Step A tten uator

Flatness" adjustmen t.

This adjustmen t zeros the digital ALC array and measures the p ow er from the RF output.

The p ow er is measured ev ery 10 MHz in low band and ev ery 100 MHz in high and millimeter

bands. The calibration constants are adjusted to ac hiev e a at output.

The calibration factors for the p ow er sensor m ust b e added to the program. Refer to the

utilit y , \4. P ow er Sensor Con guration and Calibration F actor File", if necessary .

The following test equipmen t is required for this adjustmen t. No substitutions are allo w ed.

(The test setup is given in the softw are.)

T est Equipment Required

P ow er Meter 438A

P ow er Sensor (100 kHz to 4.2 GHz) 8482A

P ow er Sensor (50 MHz to 26.5 GHz) 8485A

P ow er Sensor (50 MHz to 50 GHz)

8487A

10 dB A tten uator (DC to 26.5 GHz)

1 F or 40 GHz synthesizers only .

2 F or high p ow er synthesizers only .

Select the p ow er atness adjustmen t and en ter y our p ow er sensor data b y p erforming the

follo wing steps.

If the p ow er sensor data w as added previously , and is still curren t, skip the

following steps and run the test.

1. Select the

NNNNNNNNNNNNNN

HELP softkey to access the p ow er sensor con guration and calibration men us.

2. Select p ow er meter con guration.
3. Select the sensor to edit.
4. En ter the p ow er sensor con guration data (follow the prompts on the display).

5-14 Automated T ests Agilent 8360

Show 5 more labels / short text on this page
Description · Instrument HP/Agilent · Mo del · Numb er · Procedure
Page 412 of the Agilent Technologies 83640B user manual — RETURN
Page 412 · click to enlarge
Page 412

Do not edit the \Sensor ID". The names set at the factory m ust remain

unc hanged for the program to run prop erly . The factory recommended

\Zero Hr" is 1.00 and the \Cal Hr" is 24.00. Use the left and right arrows to

mov e the cursor within a eld and the up and do wn arrows to incremen t or

decremen t the v alue.

selects the eld.

RETURN

5. When all c hanges ha v e b een made, select sav e p ow er meter con guration.
6. Next select p ow er meter calibration factors to add the p ow er sensor calibration factors to

the program.

7. Select curren t to c hange the active p ow er sensor to the one for whic h y ou w ould like to

en ter the data.

8. Select edit header and en ter the serial n um b ers of the p ow er sensor.
9. Select edit calibration factors.

Agilent 8360 Automated T ests 5-15

Page 413 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 413 · click to enlarge
Page 413
10. En ter the calibration factors from y our p ow er sensor (follo w the prompts on the display).
11. When all of the calibration factors ha v e b een added, select store calibration factors to

store the calibration factors for future use.

12. Exit the p ow er meter calibration.
13. Run the p ow er atness adjustmen t.

5-16 Automated T ests Agilent 8360

Page 414 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 414 · click to enlarge
Page 414
1. Front P anel Emulation

This utility simulates an HP/Agilent 8360 syn thesizer front-panel k eyb oard in an instrumen t.

T o access the front panel em ulation utility , select the front panel em ulation program in the

test men u.

The front panel em ulation program cannot run front panel functions that use a p ow er meter

or prin ter. These functions cause the Agilent 8360 B- or L-series syn thesizer to act as a

con troller, whic h con icts with the computer as the con troller on the bus. Use a substitute

front panel to run these functions, or use the tests provided with this softw are.

When using the front panel em ulation program, en ter the letter directly ab ov e the caret ( ^) to

activ ate the function or k ey . Note that the k eys are case sensitive (\L" means \ALC" and \l"

means \LOCAL"). See T able 5-4 for a detailed listing of hardk ey designations.

F or n umeric en try , tw o di eren t terminator k eys are com bined on one hardk ey . The actual

terminator dep ends on the active mo de of the syn thesizer.

F or exampleWhen pulse width is the active en try ,

en ters 10 ms. When frequency

span is the active en try ,

en ters 10 kHz.

Softkeys are designated b y menukey X, where X is 1 through 5 and corresp onds to the v e

softkey p ositions b elow the display . T o activ ate the softkey , press the function k ey on the

k eyb oard that corresp onds to the n um b er X indicated b y men uk ey .

Agilent 8360 Automated T ests 5-17

Show 7 more labels / short text on this page
Utilities · Description · Procedure · 10 · k · 10 · k
Page 415 of the Agilent Technologies 83640B user manual — RECALL
Page 415 · click to enlarge
Page 415

T able 5-4. Synthesizer Hardk ey Emulation

Hardkey Program Displa y T o Activ ate

MENU SELECT

mo d m

MOD

ALC

MARKER

USER CAL

SER VICE

PRIOR

SYSTEM

SA VE

saV e V

RECALL

MENU

USER DEFINED

MENU

ASSIGN

ENTR Y

ENTR Y ON/OFF

entry o f

5-18 Automated T ests Agilent 8360

Show 34 more labels / short text on this page
Select · · · aLc L · · mrK K · · cal c · · Srv S · · Prior P · · · Rcl R · · meNu N · · menU U · · asgn a · · · GHz/dB(m) · GHz/dBm G · · MHz/usec · MHz/uS M · · kHz/msec · kHz/mS k · · Hz,sec/ENTER · Hz/ENT H
Page 416 of the Agilent Technologies 83640B user manual — FREQUENCY
Page 416 · click to enlarge
Page 416

T able 5-4. Synthesizer Hardk ey Emulation (continued)

Hardkey Program Displa y T o Activ ate

SWEEP

SWEEP TIME

SINGLE

CONT

MENU

mEn u E

FREQUENCY

CW

ST AR T

STOP

CENTER

SP AN

MENU

POWER

PO WER LEVEL

FL TNESS ON/OFF

RF ON/OFF

MENU

INSTR UMENT ST A TE

PRESET

LOCAL

lo cal l

Agilent 8360 Automated T ests 5-19

Show 31 more labels / short text on this page
Select · · Time T · · sIng I · · cOnt O · · · cW W · · stArt A · · stop o · · Cent C · · span n · · menu u · · level v · · at t · · on/oFf F · · menu e · · preset r ·
Page 417 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 417 · click to enlarge
Page 417
2. Calibration Constants

The calibration constants utilit y provides a v ariety of w ays to manipulate the calibration

constants stored in the instrumen t.

Use this utilit y to prin t out a list of the calibration constants stored in the instrumen t, to

make a bac k-up of the calibration constants, and to restore calibration constants from a

bac k ed-up le.

En ter the password to allo w access to the calibration constants b y running the front panel

em ulation program provided with this softw are. See Chapter 4 for information on accessing

the calibration constants with the password.

Y ou may need to p erform sev eral steps to reac h y our nal goal. F or example, to mov e the

calibration constants from the instrumen t's w orking memory to a disk for bac k-up, y ou m ust

rst mov e them to the computer memory , and then from computer memory to disk. T o store

the calibration constants, y ou may use the hard disk drive, Disk 2 of y our w orking disks, or a

separate data disk. When storing data on a new disk, b e sure to format the disk b efore y ou

b egin the utility .

Select the calibration constants en try in the test men u and follo w the prompts.

3. Self-T est

This utility is executed using the front panel em ulation program. A full self-test of the

syn thesizer will b e executed and the results display ed on the computer.

Occasionally , the self-tests may fail when it is run remotely or when external

cables are connected to the instrumen t. If the self-tests fail, remov e all cables

and run the self-tests from the front panel. If the self-tests pass under these

conditions, the instrumen t is op erating prop erly .

External connections to the instrumen t during self-test may not allo w the lines

to c hange conditions within the limits of the self-test's parameters.

5-20 Automated T ests Agilent 8360

Description · Procedure · Description

Page 418 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 418 · click to enlarge
Page 418
4. P o w er Sensor Configuration and Calibration F actor File

This utility is used for creating and editing the les that store p ow er sensor calibration

factors. The p ow er sensor calibration factors are used in the \Pow er Flatness" and \Step

A tten uator" p erformance tests and adjustmen t pro cedures. This data can b e stored to the

directory or disk for future use. The p ow er meter/p ow er sensor con guration utilit y is used

to de ne whic h p ow er sensors, b y serial n um b er, are actually used when the automated test

is run. Although y ou may en ter and store calibration factors for sev eral p ow er sensors, the

program will use only the p ow er sensors that y ou ha v e iden ti ed b y serial n um b er in the

con guration utilit y .

Loading Sensor Calibration F actors

The service softw are program for loading sensor calibration factors is unfriendly . F ollowing the

steps outlined should result in a successful en try .

Do not h urry or skip steps.

T o Load Sensor Cal F actors:

1. Load the 8360 Service Softw are and run.
2. A t the \T est Men u" selected one of the follo wing: \Pow er Flatness and Accuracy T est",

\Step A tten uator Flatness T est", \Pow er Flatness Adjust", or \Step A tten uator Flatness

Adjust".

3. Press the \Help" k ey (F2) when it app ears on the function k eys to get to the \Help

Men u" where the \Pmtr Con g" (Pow er Meter Con guration) and \Pmtr Cal F actors"

selections are lo cated.

4. Scroll do wn to and select [2] \Pmtr Con g". The p ow er meter con guration routine is

only used to select sensor serial n um b ers (mo del n um b ers are xed). After selecting \Pmtr

Con g" the screen will display a list of mo del and serial n um b ers the program will use

during the test.

Imp ortan t Notes for Editing the Selections:

No new mo del n um b ers can b e added. The program only recognizes the mo del n um b ers

listed and will stop if other mo del n um b ers are en tered.

Mo del n um b ers can not b e c hanged. Only the serial n um b ers can b e c hanged in this

men u. Use serial n um b ers to track di eren t sensor mo del n um b ers.

Only edit sensor serial n um b ers in this men u.

Always exit the program b y selecting [17] \Sav e Con guration" and, then, pressing

\Done". A bug in the program may erase mo del n um b ers if this is not done. If a mo del

n um b er is missing create a new w orking disk n um b er 2 using the master disk or purge

\PMTRCONF" le from the w orking disk and cop y the \PMTRCONF" le bac k to the

w orking disk from the master disk.

The frequency range of the instrumen t determined whic h sensor mo del n um b ers listed

will b e used during the test. The serial n um b er determines whic h cal factors will b e

used. The follo wing list explains whic h mo del n um b ers are used when.

Agilent 8360 Automated T ests 5-21

Description

Page 419 of the Agilent Technologies 83640B user manual — 8487D
Page 419 · click to enlarge
Page 419

F or instrumen ts with a maxim um frequency of 26.5 GHz, the follo wing sensors are used:

8485D + 10 dB pad

8485D

8485A

8482A

F or instrumen t with a maximum frequency ab ov e 26.5 GHz, the follo wing sensors are

used:

8487D + 10 dB pad

8487D

8487A

8482A

After Selection, Con tinue with Cal Sensor Pro cedures:

5. If the sensor serial n um b ers are correct, scroll do wn to [17] \Sav e P ow er Meter

Con guration", and press \Select".

6. If a di eren t sensor serial n um b er needs to b e en tered or a new sensor serial n um b er

added, scroll to the mo del n um b er to b e edited and press \Select" to get to the \Edit

Sensor #1" screen.

7. T o c hange the serial n um b er in the \Edit Sensor" screen, scroll to the Serial Num b er

selection [4] and press \Select". Once \Select" has b een pressed, a n um b er m ust b e

en tered.

8. After \Edit Sensor" is selected, \En ter T ext" should b e display ed. En ter the serial

n um b er and press \Return", then \Done". The mo del and new serial n um b er should b e

listed in the table.

9. Repeat steps 6 through 8 un til all the correct serial n um b ers are display ed.
10. No w sav e the new serial n um b ers in the P ow er Meter Con guration le b y scrolling to [17]

\Sav e P ow er Meter Con guration" and pressing \Select" then \Done" to exit the routine.

11. If the cal factors need to b e loaded for new serial n um b ers or c hanged for existing serial

n um b ers, scroll to \Pmtr Cal F actor" selection [3], and press \Select". A list of serial

n um b ers and sensor mo del n um b ers should b e display ed.

12. Scroll do wn to the serial n um b er to b e edited and press \Select". If the correct serial

n um b er is not display ed, return to the \Pmtr Con g" program and rep eat en tering the

correct serial n um b er.

When en tering cal factors for a new serial n um b er, the message \F ailed attempt to access

le . . . " will app ear. Press \Ab ort". Read the message display ed and then press \OK".

13. The \Cal F actor Edit Men u" should no w b e display ed. This men u has 6 options. Here is

an explanation of eac h selection.

[1] Edit Cal F actorsSelect \Edit Cal F actors" to edit the cal-factors for the serial

n um b er selected. After \Select" is pressed the cal factors are display ed. Scroll to the

line to b e edited and press \Select". There are four c hoices for editing the cal factors.

V alue - V alue is used to edit a cal factor. En ter the new v alue b y typing ov er the old

data. Do not use the delete k ey .

F requency - F requency is used to edit the frequency . En ter the new v alue b y typing

ov er the old data. Do not use the delete k ey .

5-22 Automated T ests Agilent 8360

Page 420 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 420 · click to enlarge
Page 420

Insert - Inserts a line in the table at the lo cation of the p ointer.

Delete - Deletes the line in the table at the lo cation of the p ointer.

[2] Edit Sensor Information: Not used.

[3] Load Cal F actorsThis selection loads the calibration factor data le for the selected

sensor form disk or the directory sp eci ed in the MSI command.

[4] Store Cal F actors: This stores the cal factors for the selected p ow er sensor to disk or

the directory sp eci ed b y the MSI command.

[5] Curren t: Displa ys the curren t sensor serial and mo del n um b er.

[6] Serial En tryThis selection is used to create a new cal factor data table for the

selected p ow er sensor.

T o en ter frequency and correction factors for the sensor, en ter the frequency , a

comma, and then the correction factor (F requency , Correction). The frequency m ust

b e en tered in GHz (.01 = 10 MHz). Correction factors m ust b e en tered in % (97.0 =

97.0%). En ter frequencies in ascending order (low est to highest).

The rst frequency p oint m ust b e b elow the start frequency the sensor is used at: F or

rst frequency p oints en ter:

8485A/D

8482A, 8485A/D

There m ust b e a calibration factor at 50 MHz. The test softw are requires this 50

MHz calibration factor for the p ow er sensor.

The last frequency p oint m ust b e greater than the highest frequency the sensor is

used at. The last frequency p oints en tered:

8487A/D 50.0 GHz

8482A >4.2 GHz

8485A/D, 8487A/D >26.5 GHz

T o end serial en try en ter \0,0".

Select \Store Cal F actors".

14. T o exit the \Cal F actor Edit Men u" press \Done".
15. If all en tries are complete, press \Done". If additional editing is required scroll to the

desired c hoice and select

After all data is en tered and \Done" has b een selected the program should con tinue

running. If a \Time Out" error o ccurs, c hec k the setup and press \Retry". If \Retry"

fails to start the program press \Ab ort" and start the program ov er.

Agilent 8360 Automated T ests 5-23

Page 421 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNN
Page 421 · click to enlarge
Page 421

The following men u c hoices are display ed:

This is the instrumen t calibration factors utility . Use it to store and retrieve instrumen t

calibration factors. The follo wing men u c hoices will allow y ou to en ter and store calibration

factors:

[1] Review CALS in Disk File

[2] Review CALS from UUT W orking CALS

[3] Store UUT W orking CALS to Disk File

[4] Load CALS F rom a Disk File to UUT W orking CALS

[5] Cop y UUT W orking CALS to UUT Protected CALS

[6] Cop y UUT Protected CALS to UUT W orking CALS

Pm tr Con g

This is the p ow er meter con guration utility . Use it to en ter serial n um b ers for the p ow er

sensors whic h are listed in the \Sensor ID" column. F or example, when the program requires

an HP/Agilent 8487D p ow er sensor, it will use the calibration factor data that is stored for

whic hev er p ow er sensor has b een listed in the \Serial" column. If HP/Agilen t 8487A/D p ow er

sensors are used to test syn thesizers with maximum stop frequencies of 26.5 GHz, en ter

their serial n um b ers for the HP/Agilent 8485A/D c hoices. The \Sensor ID" and \Name"

columns iden tify the frequency range and p ow er lev el ov er whic h the p ow er sensor is used.

The HP/Agilen t 8487A/D can substitute for an HP/Agilent 8485D; ho w ev er, the rev erse is

not true.

After y ou select a p ow er sensor, mov e the arrow to \Serial Num b er" and en ter the new serial

NNNNNNNNNNNNNN

n um b er. Then select

DONE .

When the p ow er meter con guration is correct, scroll do wn to selection 17 and sav e the

con guration data.

Pm tr Cal F actors

This is the p ow er meter calibration factor utilit y . Use it to en ter and store calibration factors

for eac h p ow er sensor. Notice that the test softw are requires calibration factors at 50 MHz

and at one frequency higher than the sp eci ed stop frequency of the syn thesizer under test.

T o edit or en ter new calibration factors, select the serial n um b er for the p ow er sensor from the

display ed list. (This list is deriv ed from the p ow er meter con guration utility and the p ow er

sensor m ust already b e en tered in the con guration.) The calibration factors whic h w ere

previously stored will b e loaded and the editing men u will b e display ed.

5-24 Automated T ests Agilent 8360

Menu Ov erview · Calco Utility

Page 422 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 422 · click to enlarge
Page 422

Agilent 8360 Menu Maps 6-1

Menu Maps

Page 423 of the Agilent Technologies 83640B user manual — 08360125.pdf
Page 423 · click to enlarge
Page 423

6-2 Menu Maps Agilent 8360

Page 424 of the Agilent Technologies 83640B user manual — ALC Menu
Page 424 · click to enlarge
Page 424
Agilent Technologies 83640B — ALC Menu diagram (page 424)
Fig. 126 · ALC Menu
Page 425 of the Agilent Technologies 83640B user manual — Frequency Menu
Page 425 · click to enlarge
Page 425
Agilent Technologies 83640B — Frequency Menu diagram (page 425)
Fig. 127 · Frequency Menu
Page 426 of the Agilent Technologies 83640B user manual — Marker Menu
Page 426 · click to enlarge
Page 426
Agilent Technologies 83640B — Marker Menu diagram (page 426)
Fig. 128 · Marker Menu
Page 427 of the Agilent Technologies 83640B user manual — Modulation Menu
Page 427 · click to enlarge
Page 427
Agilent Technologies 83640B — Modulation Menu diagram (page 427)
Fig. 129 · Modulation Menu
Page 428 of the Agilent Technologies 83640B user manual — Power Menu
Page 428 · click to enlarge
Page 428
Agilent Technologies 83640B — Power Menu diagram (page 428)
Fig. 130 · Power Menu
Page 429 of the Agilent Technologies 83640B user manual — Service Menu
Page 429 · click to enlarge
Page 429
Agilent Technologies 83640B — Service Menu diagram (page 429)
Fig. 131 · Service Menu
Page 430 of the Agilent Technologies 83640B user manual — Sweep Menu
Page 430 · click to enlarge
Page 430
Agilent Technologies 83640B — Sweep Menu diagram (page 430)
Fig. 132 · Sweep Menu
Page 431 of the Agilent Technologies 83640B user manual — System Menu
Page 431 · click to enlarge
Page 431
Agilent Technologies 83640B — System Menu diagram (page 431)
Fig. 133 · System Menu
Page 432 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 432 · click to enlarge
Page 432
Agilent Technologies 83640B — User Cal Menu diagram (page 432)
Fig. 134 · User Cal Menu
Page 433 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 433 · click to enlarge
Page 433

Disassembly and Replacement Procedures

This c hapter provides the following disassembly and reassembly pro cedures. Use these

pro cedures while repairing or replacing an assembly .

\F ront P anel Disassembly and Reassem bly"

\Rear P anel Disassembly and Reassem bly"

\RF Dec k Disassembly and Reassem bly"

\Motherb oard Disassem bly and Reassem bly"

Agilent 8360 Disassembly and Replacement Procedures 7-1

Introduction

Page 434 of the Agilent Technologies 83640B user manual — CA UTION
Page 434 · click to enlarge
Page 434

W ARNING

These servicing instructions are for use b y qualified personnel only . T o a v oid

electrical shock, do not perform any servicing unless y ou are qualified to do

so.

The opening of co v ers or remo v al of parts is lik ely to expose dangerous

v oltages. Disconnect the instrument from all v oltage sources while it is being

opened.

The detachable po w er cord is the instrument disconnecting device. It

disconnects the mains circuits from the mains supply before other parts of

the instrument. The front panel switch is only a standb y switch and is not a

LINE switch.

The po w er cord is connected to internal capacitors that may remain liv e for 5

seconds after disconnecting the plug from its po w er supply .

This is a Safety Class I product (pro vided with a protectiv e earthing ground

incorporated in the po w er cord). The mains plug shall only be inserted in a

sock et outlet pro vided with a protectiv e earth contact. Any interruption of the

protectiv e conductor, inside or outside the instrument, is lik ely to mak e the

instrument dangerous. Intentional interruption is prohibited.

For continued protection against fire hazard replace line fuse only with same

type and rating (F 5A/250V). The use of other fuses or material is prohibited.

CA UTION

P erform the following pro cedures only at a static safe w ork station. The

attenuators and prin ted circuit assemblies in this instrumen t are v ery

sensitive to static electricit y damage. W ear an an ti-static wrist strap that is

connected to earth ground.

When reassembling the instrumen t, the cables sho wn in Figure 7-1 m ust b e

routed through the instrumen t as shown or instrumen t p erformance may b e

a ected.

7-2 Disassembly and Replacement Procedures Agilent 8360

Page 435 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 435 · click to enlarge
Page 435
Agilent Technologies 83640B — User Cal Menu diagram (page 435)
Fig. 135 · User Cal Menu

Figure 7-1. Cable Routing Locations

Eac h pro cedure lists the required to ols at the b eginning of that section. See Chapter 9 for the

asso ciated part n um b ers.

Agilent 8360 Disassembly and Replacement Procedures 7-3

T ools Required

Page 436 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 436 · click to enlarge
Page 436

Front P anel Disassembly and Reassembly

T-10 TORX Screwdriv er (torque 10 in-lb)

0.050 Hex Driver

9/16 in. Nut Driv er

5/16 in. Op en End W renc h

Coax Extractor T o ol{Submin D Connector

Some steps in this pro cedure may not apply to all option con gurations. If a particular step

do es not apply , con tinue with the next appropriate step.

T o disassemble the front panel, follow this pro cedure in the order given.

T o reassemble the front panel, follo w this pro cedure in the rev erse order.

Refer to the o w c hart in Figure 7-2 to determine the necessary disassembly steps for a given

assembly .

7-4 Disassembly and Replacement Procedures

T ools Required · Description and Procedure · Front P anel

Page 437 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 437 · click to enlarge
Page 437
Agilent Technologies 83640B — User Cal Menu diagram (page 437)
Fig. 136 · User Cal Menu

Figure 7-2. Front P anel Disassembly Flo w chart

Agilent 8360 Disassembly and Replacement Procedures

Front P anel

Page 438 of the Agilent Technologies 83640B user manual — CA UTION
Page 438 · click to enlarge
Page 438
1. Preliminary Steps

CA UTION

Y ou can easily blow the fuse for the 13.5 V supply on the preregulator

assembly if y ou do not disconnect the A C p ow er cord. Also, the danger of

sho c k is increased if the p ow er cord is connected.

1. Disconnect the A C p ow er line cord.
2. Remo v e the screw from eac h of the top tw o feet on the rear panel.
3. Remo v e the top tw o feet.
4. Remo v e the instrumen t top co v er.

An RF braid is in eac h of the side c hannels on the top of the instrumen t.

When y ou reassemble the instrumen t, b e sure this braid remains in the

c hannel.

5. Remo v e the vin yl trim strip from the top front edge of the instrumen t. (Place a at

screwdriver in either slot in the trim strip and lift to remov e, see Figure 7-3.)

Agilent Technologies 83640B — CA UTION diagram (page 438)
Fig. 137

Figure 7-3. Front Edge Trim Strip

7-6 Disassembly and Replacement Procedures

Front P anel

Page 439 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 439 · click to enlarge
Page 439
2. Front P anel Remov al

Refer to Figure 7-4.

1. Complete \1. Preliminary Steps" b efore con tinuing.
2. Remo v e thr e e of the four screws under the trim strip.
3. Remo v e two of the three screws on the low er front frame edge.
4. Gen tly pry the front panel and display from the frame.

If y ou are removing the front panel only so that y ou can lift up the RF dec k

and if y ou ha v e a at surface in front of the syn thesizer on whic h to lay the

front panel, let the front panel lie in front of the syn thesizer with all the cables

connected and con tinue with step 1 of \2. Lift Up the RF Dec k". If y ou do

not ha v e a at surface for the front panel, disconnect ribb on cables W2 and

W3 from the motherb oard and lay the front panel on top of the syn thesizer.

Agilent Technologies 83640B — User Cal Menu diagram (page 439)
Fig. 138 · User Cal Menu

Figure 7-4. Front P anel Screw Locations

Agilent 8360 Disassembly and Replacement Procedures

Front P anel

Page 440 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 440 · click to enlarge
Page 440

Refer to Figure 7-5.

Instrumen ts with Option 004 do not ha v e all the cables sho wn.

5. Cut the tie wraps holding the cables.
6. F rom the A9, A10, and A11 assemblies (or the A8 assembly for instrumen ts with Option

002), disconnect the four coaxial cables (W6, W7, W8, and W9) attached to the BNC

connectors.

7. Disconnect the source mo dule in terface cable, W10.
8. Disconnect ribb on cables W2 and W3 from the motherb oard.
Agilent Technologies 83640B — User Cal Menu diagram (page 440)
Fig. 139 · User Cal Menu

Figure 7-5. Front P anel Coaxial and Ribbon Cable Locations

7-8 Disassembly and Replacement Procedures

Front P anel

Page 441 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 441 · click to enlarge
Page 441
3. A3 Front P anel Processor Remov al

Refer to Figure 7-6.

1. Complete \1. Preliminary Steps" and \2. F ront P anel Remo v al" b efore con tinuing.
2. Disconnect the three ribb on cables (W3, W4, and W5), the rotary pulse generator (RPG)

cable, and the six-wire display cable W1.

3. Remo v e the eigh t screws from the PC assembly .
4. Lift o the front panel pro cessor assembly .
Agilent Technologies 83640B — User Cal Menu diagram (page 441)
Fig. 140 · User Cal Menu

Figure 7-6. Front P anel Processor Cables

Agilent 8360 Disassembly and Replacement Procedures

Front P anel

Page 442 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 442 · click to enlarge
Page 442
4. A21 Alphanumeric Displa y Remov al

Refer to Figure 7-7.

1. Complete \1. Preliminary Steps" and \2. F ront P anel Remo v al" b efore con tinuing.
2. Disconnect the six-wire display cable W1 and ribb on cable W4.
3. Remo v e the four outer-most screws around the edge of the brac k et.
4. Lift o the display .
Agilent Technologies 83640B — User Cal Menu diagram (page 442)
Fig. 141 · User Cal Menu

Figure 7-7. Display Remo v al

7-10 Disassembly and Replacement Procedures

Front P anel

Page 443 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 443 · click to enlarge
Page 443
5. A2 Source Module Interface Remov al

Refer to Figure 7-8.

1. Complete \1. Preliminary Steps", \2. F ront P anel Remo v al", and \3. A3 F ront P anel

Pro cessor Remo v al" b efore con tinuing.

2. Remo v e the source mo dule in terface cable (W10).
3. Remo v e the four screws.
4. Remo v e the source mo dule in terface assembly .
5. Disconnect ribb on cable W2 from the source mo dule in terface assembly .
6. Remo v e the A2 source mo dule in terface assembly .
Agilent Technologies 83640B — User Cal Menu diagram (page 443)
Fig. 142 · User Cal Menu

Figure 7-8. Source Module Interface Remo v al

Agilent 8360 Disassembly and Replacement Procedures

Front P anel

Page 444 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 444 · click to enlarge
Page 444
6. RPG1 Rotary Pulse Generator Remov al

Refer to Figure 7-9.

1. Complete \1. Preliminary Steps", \2. F ront P anel Remo v al", and \3. A3 F ront P anel

Pro cessor Remo v al" b efore con tinuing.

2. Lo osen the tw o hex screws and remov e the rotary knob.
Agilent Technologies 83640B — User Cal Menu diagram (page 444)
Fig. 143 · User Cal Menu

Figure 7-9. RPG Remo v al

7-12 Disassembly and Replacement Procedures

Front P anel

Page 445 of the Agilent Technologies 83640B user manual — CA UTION
Page 445 · click to enlarge
Page 445
7. J1 RF Output Connector Remov al

Refer to Figure 7-10 for front panel output.

CA UTION

T o av oid damaging the connector, do not disassemble the RF output

connector assembly . Only unscrew the connector on the end of the cable.

1. Complete \1. Preliminary Steps" and \2. F ront P anel Remo v al" b efore con tinuing.
2. On the low er front frame edge, remov e the tw o RF output screws that hold the RF output

assembly .

3. Remo v e the RF connector with cable W43.
Agilent Technologies 83640B — CA UTION diagram (page 445)
Fig. 144

Figure 7-10. Remo ving RF Output Connector (Front P anel)

When y ou re-attach the connector and cable to the syn thesizer, rst, lo osen

the n ut b et w een cable W43 and the connector. Then rev erse steps 1 and 2.

Finally , tighten the n ut b et w een cable W43 and the connector to 10 in-lb.

Agilent 8360 Disassembly and Replacement Procedures

Front P anel

Page 446 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 446 · click to enlarge
Page 446
8. Front P anel BNC Connectors Remov al

Refer to Figure 7-11.

1. Complete \1. Preliminary Steps", \2. F ront P anel Remo v al", \3. A3 F ront P anel Pro cessor

Remo v al" and \5. A2 Source Mo dule In terface Remo v al" b efore con tinuing.

2. Remo v e the tw o screws on eac h of the four PC assemblies attached to the bac k of the front

panel.

3. Slide out the connectors.

If the w asher on the front of the connector is not cen tered prop erly , lo osen the

n ut to slide the connector out.

Agilent Technologies 83640B — User Cal Menu diagram (page 446)
Fig. 145 · User Cal Menu

Figure 7-11. Front P anel BNC Locations

7-14 Disassembly and Replacement Procedures

Front P anel

Page 447 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 447 · click to enlarge
Page 447
9. A1 Keyboard Remov al

Refer to Figure 7-12.

1. Complete \1. Preliminary Steps", \2. F ront P anel Remo v al", \3. A3 F ront P anel Pro cessor

Remo v al", \4. A21 Alphan umeric Display Remo v al" and \5. A2 Source Mo dule In terface

Remo v al" b efore con tinuing.

2. Remo v e the nine screws from the k eyb oard assembly .
3. Lift o the k eyb oard assembly .

It is imp ortant that the top of the k eyb oard b e k ept clean. Dirt on the

con tacts will a ect the p erformance of the k eypad.

4. Disconnect ribb on cable W5 from the k eyb oard.
Agilent Technologies 83640B — User Cal Menu diagram (page 447)
Fig. 146 · User Cal Menu

Figure 7-12. Keyboard Assembly Remo v al

Refer to Figure 7-13.

Agilent 8360 Disassembly and Replacement Procedures

Front P anel

Page 448 of the Agilent Technologies 83640B user manual — CA UTION
Page 448 · click to enlarge
Page 448

CA UTION

The follo wing steps can degrade k eypad p erformance. If y ou separate the

an ti-ro c k sheets from the k eypad, y ou m ust replace b oth the an ti-ro c k sheets

and the k eypad.

5. Remo v e the k eypad.
6. Remo v e the an ti-ro c k sheets.
Agilent Technologies 83640B — CA UTION diagram (page 448)
Fig. 147

Figure 7-13. Keypad and Anti-rock Sheets

7-16 Disassembly and Replacement Procedures

Front P anel

Page 449 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 449 · click to enlarge
Page 449
10. Displa y Filter Remov al

Refer to Figure 7-14.

1. Complete \1. Preliminary Steps", \2. F ront P anel Remo v al", \3. A3 F ront P anel Pro cessor

Remo v al", \4. A21 Alphan umeric Display Remo v al", \5. A2 Source Mo dule In terface

Remo v al" and \9. A1 Keyb oard Remo v al" b efore con tinuing.

2. With all other assemblies remov ed, lift out the display lter.
Agilent Technologies 83640B — User Cal Menu diagram (page 449)
Fig. 148 · User Cal Menu

Figure 7-14. Display Filter

Agilent 8360 Disassembly and Replacement Procedures

Front P anel

Page 450 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 450 · click to enlarge
Page 450
11. A1 Keyboard Reassembly

Refer to Figure 7-15.

1. With the front panel frame supp orted on the edges so it is elev ated o the table b y at least

one inc h, place the frame face do wn.

2. Clean the display lter and place it on the frame.
3. Carefully lining up the presst holes, place the an ti-ro c k sheets ov er the k eypad. Press

in to place.

4. Place the k eypad face do wn and press eac h k ey in to place.
5. Insp ect the front panel to b e sure all k eys protrude through the front panel, and are all the

same heigh t.

Agilent Technologies 83640B — User Cal Menu diagram (page 450)
Fig. 149 · User Cal Menu

Figure 7-15. Keyboard Reassembly

Refer to Figure 7-16

6. Connect ribb on cable W5 to the A1 k eyb oard assembly .
7. Replace the A1 k eyb oard assembly .

7-18 Disassembly and Replacement Procedures

Front P anel

Page 451 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 451 · click to enlarge
Page 451
8. Replace the nine screws that hold in the assembly .
Agilent Technologies 83640B — User Cal Menu diagram (page 451)
Fig. 150 · User Cal Menu

Figure 7-16. Keyboard Assembly Screw Locations

Agilent 8360 Disassembly and Replacement Procedures

Front P anel

Page 452 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 452 · click to enlarge
Page 452
12. Front P anel Reassemblies

F ollo w the o w c hart, Figure 7-2, in rev erse order to p erform the reassembly of the follo wing

assemblies:

J1 RF Output Connector

A3 F ront P anel Pro cessor

A21 Alphan umeric Display

A2 Source Mo dule In terface Assem bly

RPG1 Rotary Pulse Generator (RPG)

7-20 Disassembly and Replacement Procedures

BNC Connectors · Displa y Filter · Front P anel

Page 453 of the Agilent Technologies 83640B user manual — CA UTION
Page 453 · click to enlarge
Page 453

Rear P anel Disassembly and Reassembly

9/16 in. So c k et

T-10 T orx Screwdriv er

T-15 T orx Screwdriv er

5.5 mm Nut Driver

5/16 in. Op en End W renc h

14 mm Op en End W renc h

7 mm Nut Driver

#1 x 3 1/8 in. P ozidrive Screwdriv er

#2 x 4 in. P ozidriv Screwdriv er

Coax Extractor T o ol{Submin D Connector

Some steps in this pro cedure may not apply to all option con gurations. If a particular step

do es not apply , con tinue with the next appropriate step.

T o disassemble the rear panel and asso ciated assemblies, follo w this pro cedure in the order

given.

T o reassemble the rear panel and asso ciated assemblies, follo w this pro cedure in the rev erse

order.

CA UTION

Cable routing may a ect instrumen t p erformance. Refer to Figure 7-1 for a

diagram of the instrumen t cable routing.

Refer to the o w c hart in Figure 7-17 to determine the necessary disassembly steps for a given

assembly .

Agilent 8360 Disassembly and Replacement Procedures

T ools Required · Description and Procedure · Rear P anel

Page 454 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 454 · click to enlarge
Page 454
Agilent Technologies 83640B — User Cal Menu diagram (page 454)
Fig. 151 · User Cal Menu

Figure 7-17. Rear P anel Disassembly Flo w chart

7-22 Disassembly and Replacement Procedures

Rear P anel

Page 455 of the Agilent Technologies 83640B user manual — CA UTION
Page 455 · click to enlarge
Page 455
1. Preliminary Steps

CA UTION

Y ou can easily blow the fuse for the 13.5 V supply on the preregulator

assembly if y ou do not disconnect the A C p ow er cord. Also, the danger of

sho c k is increased if the p ow er cord is connected.

1. Disconnect the A C p ow er line cord.
2. Remo v e the screw from eac h of the four feet on the rear panel.
3. Remo v e the four feet.
4. T o remov e the side straps, remov e the tw o screws on eac h side panel.
5. Remo v e the instrumen t top, b ottom, and side co v ers.

An RF braid is in eac h of the side c hannels on the top and b ottom of the

instrumen t. When y ou reassemble the instrumen t, b e sure the braids remain

in the side co v er c hannels.

Agilent 8360 Disassembly and Replacement Procedures

Rear P anel

Page 456 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 456 · click to enlarge
Page 456
2. B1 F an Remov al

Refer to Figure 7-18.

1. Complete \1. Preliminary Steps" b efore con tinuing.
2. Remo v e b oth screws (item 1) from the fan shroud.
3. Remo v e items 2 through 5.
4. Remo v e the four screws (item 7) from the rear panel/fan brac k ets.
5. Unplug the fan harness.
6. Tip the fan and angle it out through the op ening in the rear panel.
7. Remo v e the four screws (item 7) attaching the tw o brac k ets to the fan and remov e the

brac k ets (item 10).

Agilent Technologies 83640B — User Cal Menu diagram (page 456)
Fig. 152 · User Cal Menu

Figure 7-18. B1 Rear P anel F an Assembly

7-24 Disassembly and Replacement Procedures

Rear P anel

Page 457 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 457 · click to enlarge
Page 457
3. Rear P anel BNC Connectors Remov al and Aux Output

Refer to Figure 7-19.

1. Complete \1. Preliminary Steps" b efore con tinuing.
2. Remo v e the n uts and w ashers on the four BNC connectors, 10 MHz REF INPUT, 10 MHz

REF OUTPUT, V OL TS/GHz, and TRIGGER OUTPUT.

3. Slide out the four connectors.
4. If y ou ha v e Option 002, rep eat steps 1 and 2 for the following connectors: AM/FM

OUTPUT, PULSE SYNC OUT, and PULSE VIDEO OUT.

5. Disconnect W48 at rear panel SMA connector J14 A UX OUTPUT.
Agilent Technologies 83640B — User Cal Menu diagram (page 457)
Fig. 153 · User Cal Menu

Figure 7-19. Rear P anel BNC Connector Locations

Agilent 8360 Disassembly and Replacement Procedures

Disconnection · Rear P anel

Page 458 of the Agilent Technologies 83640B user manual — CA UTION
Page 458 · click to enlarge
Page 458
4. J1 RF Output Connector Remov al (Option 004)

Refer to Figure 7-20.

CA UTION

T o av oid damaging the connector, do not disassemble the RF output

connector assembly . Only unscrew the connector on the end of the cable.

1. Complete \1. Preliminary Steps" b efore con tinuing.
2. Disconnect cable W43 at the directional coupler A30 (or at the step attenuator A31 for

instrumen ts with Option 001).

3. Remo v e the tw o screws from the RF output connector plate on the rear panel.
4. Remo v e the RF connector with cable W43 attached.
Agilent Technologies 83640B — CA UTION diagram (page 458)
Fig. 154

Figure 7-20. Remo ving RF Output Connector

When y ou re-attach the connector and cable to the syn thesizer, rst, lo osen

the n ut b et w een cable W43 and the connector. Then p erform the disassembly

steps in rev erse order. Finally , torque the n ut b et w een cable W43 and the

connector J1 to 10 in-lb (112 N-cm).

7-26 Disassembly and Replacement Procedures

Rear P anel

Page 459 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 459 · click to enlarge
Page 459
5. Rear P anel Remov al

Refer to Figure 7-21.

1. Complete \1. Preliminary Steps" and \7. J1 RF Output Connector Remo v al" b efore

con tinuing.

2. Remo v e the four screws on the top bac k edge.
3. Remo v e the four screws on the b ottom bac k edge.
4. Slide the rear panel out of the casting.
Agilent Technologies 83640B — User Cal Menu diagram (page 459)
Fig. 155 · User Cal Menu

Figure 7-21. Rear P anel Remo v al

Agilent 8360 Disassembly and Replacement Procedures

Rear P anel

Page 460 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 460 · click to enlarge
Page 460

Refer to Figure 7-22.

5. Disconnect ribb on cable W31 from the motherb oard.
6. Disconnect the 8-pin line switch connector.
7. Disconnect the fan harness.
8. Cut the tie wraps holding the coaxial cables to the four BNC connectors and the source

mo dule in terface.

9. Disconnect the coaxial cables from the A7, A12, and A14 assemblies and disconnect the

source mo dule in terface cable from the A19 assembly .

Agilent Technologies 83640B — User Cal Menu diagram (page 460)
Fig. 156 · User Cal Menu

Figure 7-22. Rear P anel Cable Locations

7-28 Disassembly and Replacement Procedures

Rear P anel

Page 461 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 461 · click to enlarge
Page 461
6. A23 10 MHz Reference Standard Remov al

Refer to Figure 7-23.

1. Complete \1. Preliminary Steps", \7. J1 RF Output Connector Remo v al", and \5. Rear

P anel Remo v al" b efore con tinuing.

2. Disconnect cable W17 and the pin-lo c king tab, W30.
3. Without removing the screws, press the sho c k-moun t pads through the sheet metal holes.
4. Slide the reference standard out.
Agilent Technologies 83640B — User Cal Menu diagram (page 461)
Fig. 157 · User Cal Menu

Figure 7-23. Reference Standard Mountings

Agilent 8360 Disassembly and Replacement Procedures

Rear P anel

Page 462 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 462 · click to enlarge
Page 462
7. T1 Transformer Remov al

Refer to Figure 7-24.

1. Complete \1. Preliminary Steps", \7. J1 RF Output Connector Remo v al", and \5. Rear

P anel Remo v al" b efore con tinuing.

2. Remo v e the six-wire pin-lo c king tab.
3. Remo v e the cen ter screw, plastic w asher, tw o insulators, and plate.
4. Remo v e the toroidal transformer.
Agilent Technologies 83640B — User Cal Menu diagram (page 462)
Fig. 158 · User Cal Menu

Figure 7-24. Transformer Remo v al

7-30 Disassembly and Replacement Procedures

Rear P anel

Page 463 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 463 · click to enlarge
Page 463
8. A19 Rear P anel Interface Remov al

Refer to Figure 7-25.

1. Complete \1. Preliminary Steps", \7. J1 RF Output Connector Remo v al", and \5. Rear

P anel Remo v al" b efore con tinuing.

2. Disconnect the source mo dule in terface cable W23 from SMB connector A10J3 on the A10

ALC assembly . (W23 is not included with replacemen t A19 assemblies.)

3. Remo v e the v e screws.
4. Remo v e A19 from the rear panel.
5. T o remov e the shield from the rear panel in terface, rst remov e W23 using the coax

extractor to ol and then remov e the tw o screws holding the shield to the rear panel

in terface.

Agilent Technologies 83640B — User Cal Menu diagram (page 463)
Fig. 159 · User Cal Menu

Figure 7-25. A19 Rear P anel Interface Remo v al

Agilent 8360 Disassembly and Replacement Procedures

Rear P anel

Page 464 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 464 · click to enlarge
Page 464
9. V oltage Selector Switch Remov al

Refer to Figure 7-26.

1. Complete \1. Preliminary Steps", \7. J1 RF Output Connector Remo v al", and \5. Rear

P anel Remo v al" b efore con tinuing.

2. Cut the tie wraps holding the wires.
3. Disconnect the 8-pin connector from the motherb oard.
4. Remo v e the tw o n uts, b olts, and w ashers holding in the v oltage selector switch.
Agilent Technologies 83640B — User Cal Menu diagram (page 464)
Fig. 160 · User Cal Menu

Figure 7-26. V oltage Selection Switch Remo v al

7-32 Disassembly and Replacement Procedures

Rear P anel

Page 465 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 465 · click to enlarge
Page 465

Refer to Figure 7-27.

5. Unsolder the white/brown/grey wire from the fuse.
6. Unsolder the white/grey wire from the line lter.
7. Unscrew the green/white ground wire from the bac k panel.
Agilent Technologies 83640B — User Cal Menu diagram (page 465)
Fig. 161 · User Cal Menu

Figure 7-27. Fuse, Line Filter, and Ground Wires

T o reassemble, add 3/8 in. shrink tubing to eac h wire b efore soldering. Then

p erform the ab ov e steps in rev erse order.

Agilent 8360 Disassembly and Replacement Procedures

Rear P anel

Page 466 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 466 · click to enlarge
Page 466
10. Fuse Housing Remov al

Refer to Figure 7-28.

1. Complete \1. Preliminary Steps", \7. J1 RF Output Connector Remo v al", and \5. Rear

P anel Remo v al" b efore con tinuing.

2. Remo v e the tw o n uts, b olts, and w ashers holding the v oltage selector switch, and remov e

the switch (this provides clearance for the wrenc h to remov e the fuse holder).

3. Unsolder the tw o fuse holder wires.
4. Unscrew the n ut holding the fuse housing.
5. Slide the fuse housing out of the bac k panel.
Agilent Technologies 83640B — User Cal Menu diagram (page 466)
Fig. 162 · User Cal Menu

Figure 7-28. Rear P anel Fuse Housing

T o reassemble, add 3/8 in. shrink tubing to eac h wire b efore soldering. Then

p erform the ab ov e steps in rev erse order.

7-34 Disassembly and Replacement Procedures

Rear P anel

Page 467 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 467 · click to enlarge
Page 467
11. FL1 Line Filter Remov al

Refer to Figure 7-29.

1. Complete \1. Preliminary Steps", \7. J1 RF Output Connector Remo v al", and \5. Rear

P anel Remo v al" b efore con tinuing.

2. Unsolder the three wires attached to the line lter.
3. Remo v e the tw o screws on the rear panel.
4. Slide out the line lter from the rear panel.
Agilent Technologies 83640B — User Cal Menu diagram (page 467)
Fig. 163 · User Cal Menu

Figure 7-29. Line Filter Remo v al

T o reassemble, add 3/8 in. shrink tubing to eac h wire b efore soldering. Then

p erform the ab ov e steps in rev erse order.

Agilent 8360 Disassembly and Replacement Procedures

Rear P anel

Page 468 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 468 · click to enlarge
Page 468
12. Option 004 BNC Connectors Remov al

Refer to Figure 7-30.

1. Complete \1. Preliminary Steps" b efore con tinuing.
2. Remo v e the v e screws holding the A34 RP oating BNC b oard assembly to the rear

panel.

3. Slide out the b oard with the connectors.
4. Remo v e the n ut and w asher attac hing the pulse input BNC connector to the rear panel.
5. Slide out the pulse input BNC connector.
Agilent Technologies 83640B — User Cal Menu diagram (page 468)
Fig. 164 · User Cal Menu

Figure 7-30. Option 004 BNC Connectors Remo v al

7-36 Disassembly and Replacement Procedures

Rear P anel

Page 469 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 469 · click to enlarge
Page 469

RF Deck Disassembly and Reassembly

#1 x 3 1/8 in. P ozidriv Screwdriv er

T-10 T orx Screwdriv er

5/16 in. T orque W renc h

5 mm Ball and So c k et Driver

On the RF dec k, all P ozidriv screws ha v e English threads and all T orx screws

ha v e metric threads.

T o disassemble the RF dec k and asso ciated assemblies, follo w this pro cedure in the order

given.

T o reassemble the RF dec k and asso ciated assemblies, follo w this pro cedure in the rev erse

order.

Refer to the o w c hart in Figure 7-31 to determine the necessary disassembly steps for a given

assembly .

Agilent 8360 Disassembly and Replacement Procedures

T ools Required · Description and Procedure · RF Deck

Page 470 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 470 · click to enlarge
Page 470
Agilent Technologies 83640B — User Cal Menu diagram (page 470)
Fig. 165 · User Cal Menu

Figure 7-31. RF Deck Disassembly Flo w chart

7-38 Disassembly and Replacement Procedures

RF Deck

Page 471 of the Agilent Technologies 83640B user manual — CA UTION
Page 471 · click to enlarge
Page 471
1. Preliminary Steps

CA UTION

Y ou can easily blow the fuse for the 13.5 V supply on the preregulator

assembly if y ou do not disconnect the A C p ow er cord. Also, the danger of

sho c k is increased if the p ow er cord is connected.

1. Disconnect the A C p ow er line cord.
2. Remo v e the screw from eac h of the top tw o feet on the rear panel.
3. Remo v e the top tw o feet.
4. Remo v e the instrumen t top co v er.

An RF braid is in eac h of the side c hannels on the top of the instrumen t.

When y ou reassemble the instrumen t, b e sure this braid remains in the

c hannel.

2. Lift Up the RF Deck

If y our instrumen t has front panel RF output, remov e the front panel. (See \F ront P anel

Disassembly and Reassem bly", step 2.)

If y our instrumen t has rear panel RF output (Option 004), see step 4 of \Rear P anel

Disassembly and Reassem bly" then con tinue with step 4 of this pro cedure.

CA UTION

T o av oid damaging the connector, do not disassemble the RF output

connector assembly . Only unscrew the connector on the end of the cable.

1. Complete \1. Preliminary Steps" b efore con tinuing.
2. On the low er front frame edge, remov e the tw o screws that hold the RF output assembly .
3. Remo v e the RF connector with cable W43 still connected.

Agilent 8360 Disassembly and Replacement Procedures

RF Deck

Page 472 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 472 · click to enlarge
Page 472
Agilent Technologies 83640B — User Cal Menu diagram (page 472)
Fig. 166 · User Cal Menu

Figure 7-32. Remo ving RF Output Connector

When y ou re-attach the connector and cable to the syn thesizer, rst lo osen the

n ut b et w een cable W43 and the connector. Then rev erse steps 1 through 3.

Finally , torque the n ut b et w een cable W43 and connector J1 to 10 in-lb (112

N-cm).

4. Reinstall the front panel.
5. Carefully disconnect cable W48 from the rear panel SMA connector W14 (it's a close t {

w atch out that y ou do not damage the cable cen ter conductor).

6. Disconnect cable W51 (see Figure 7-33) from the low pass lter FL2

(you may need to lo osen the other end of W51 also).

7. Remo v e the tw o hex screws from the RF dec k.

7-40 Disassembly and Replacement Procedures

RF Deck

Page 473 of the Agilent Technologies 83640B user manual — CA UTION
Page 473 · click to enlarge
Page 473
Agilent Technologies 83640B — User Cal Menu diagram (page 473)
Fig. 167 · User Cal Menu

Figure 7-33. W51 RF Cable Location

CA UTION

Av oid damage to the v arious ribb on cables and their connectors attached to

the RF dec k assembly when removing that assembly in the next step. Mov e

all exible coaxial cables out of the w ay of the RF dec k assembly comp onen ts.

8. Carefully pull out the RF dec k and rest it on the side of the instrumen t.

Some exible cables will still b e attached prev en ting the RF dec k from b eing

completely remov ed from the instrumen t.

Refer to Figure 7-34.

The analog extender b oard and coax extender cable men tioned in the follo wing

steps are part of the service to ol kit. Refer to Chapter 9 for the part n um b er

of this kit.

Agilent 8360 Disassembly and Replacement Procedures

RF Deck

Page 474 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 474 · click to enlarge
Page 474
9. F or troublesho oting insert an analog extender b oard in to the motherb oard connector for

the RF dec k. Place the RF dec k on the extender b oard and use the tw o hex screws to

attach the dec k to the syn thesizer frame (service p osition). If y ou are not troublesho oting

and are lifting up the RF dec k for disassembly purp oses, attac h the dec k to the

syn thesizer frame without using the extender b oard.

10. F or troublesho oting:

a. A ttac h a coax extender cable b et w een W51 and the low pass lter FL2.

b. Reconnect the RF connector and W43.

i. Lo osen the n ut b et w een cable W43 and the connector.

ii. Reconnect cable W43 at directional coupler A30 (or the attenuator A31 for

instrumen ts with Option 001).

iii. T orque the n ut b et w een cable W43 and connector J1 to 10 in-lb (112 N-cm).

Agilent Technologies 83640B — User Cal Menu diagram (page 474)
Fig. 168 · User Cal Menu

Figure 7-34. RF Deck in the Service P osition

7-42 Disassembly and Replacement Procedures

RF Deck

Page 475 of the Agilent Technologies 83640B user manual — W41
Page 475 · click to enlarge
Page 475
3. A20 RF Interface Remov al

Refer to Figure 7-35.

1. Complete \1. Preliminary Steps" and \2. Lift Up the RF Dec k" b efore con tinuing.
2. Disconnect the follo wing ribb on cables. (Y ou may need to lo osen the tw o hex screws

attaching the dec k to the syn thesizer frame in order to remov e the cables on the low er edge

of the RF dec k.)

W36 ( 83640B /L and 83650B /L only)

W37

W38

W39

W40 (Option 001 only)

W41

W42

W69 (All mo dels except 83624B )

W73 (All mo dels except 83623B /L and 83624B )

3. Remo v e the 3-wire pin-lo c king tab ( 83623B /L and 83624B ).
4. Remo v e the four corner screws that attach the b oard to the casting.
5. Remo v e the A20 RF in terface assembly .
Agilent Technologies 83640B — W41 diagram (page 475)
Fig. 169

Figure 7-35. A20 RF Interface

Agilent 8360 Disassembly and Replacement Procedures

RF Deck

Page 476 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 476 · click to enlarge
Page 476
4. Microcircuit Remov al

F or the follo wing pro cedures, refer to Figure 7-36 through Figure 7-50 for

cable lo cations and for screw lo cations.

1. Complete \1. Preliminary Steps" and \2. Lift Up the RF Dec k" b efore con tinuing.
2. With the RF dec k up in the service p osition, an y micro circuit can b e remov ed:

a. Most micro circuits ha v e a cable connecting them to the A20 RF in terface assembly .

Disconnect the cable from the A20 assembly (if y ou ha v e already remov ed the A20

assembly , this step is not necessary).

b. Disconnect all semi-rigid and exible cables from the micro circuit. Be sure to lo osen

all semi-rigid cables at their other connections also. If y ou do not, the cables may b e

damaged.

c. Remo v e the micro circuits b y removing the screws that hold them to the RF dec k.

7-44 Disassembly and Replacement Procedures

RF Deck

Page 477 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 477 · click to enlarge
Page 477
Agilent Technologies 83640B — User Cal Menu diagram (page 477)
Fig. 170 · User Cal Menu

Figure 7-36. RF Deck Cable Locations { 83640B/L and 83650B/L

Agilent 8360 Disassembly and Replacement Procedures

Show 4 more labels / short text on this page
RF Deck Cables · 83640B/L and 83650B/L · No Options · RF Deck
Page 478 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 478 · click to enlarge
Page 478
Agilent Technologies 83640B — User Cal Menu diagram (page 478)
Fig. 171 · User Cal Menu

Figure 7-37.

RF Deck Cable Locations {

83640B/L Option 001 and 83650B/L Option 001

7-46 Disassembly and Replacement Procedures

Show 4 more labels / short text on this page
RF Deck Cables · 83640B/L and 83650B/L · Option 001 · RF Deck
Page 479 of the Agilent Technologies 83640B user manual — 83640B/50B
Page 479 · click to enlarge
Page 479

83640B/50B

Agilent Technologies 83640B — 83640B/50B diagram (page 479)
Fig. 172

Figure 7-38. RF Deck Cable Locations { 83640B/50B Option 006

Agilent 8360 Disassembly and Replacement Procedures

RF Deck Cables · Option 006 · RF Deck

Page 480 of the Agilent Technologies 83640B user manual — 83640B/50B;
Page 480 · click to enlarge
Page 480

83640B/50B;

Options 001 and 006

Agilent Technologies 83640B — 83640B/50B; diagram (page 480)
Fig. 173

Figure 7-39. RF Deck Cable Locations { 83640B/50B Options 001 and 006

7-48 Disassembly and Replacement Procedures

RF Deck Cables · RF Deck

Page 481 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 481 · click to enlarge
Page 481
Agilent Technologies 83640B — User Cal Menu diagram (page 481)
Fig. 174 · User Cal Menu

Figure 7-40. RF Deck Cable Locations { 83620B/22B and 83630B/L

Agilent 8360 Disassembly and Replacement Procedures

Show 4 more labels / short text on this page
RF Deck Cables · 83620B/22B and 83630B/L · No Options · RF Deck
Page 482 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 482 · click to enlarge
Page 482
Agilent Technologies 83640B — User Cal Menu diagram (page 482)
Fig. 175 · User Cal Menu

Figure 7-41. RF Deck Cable Locations { 83620B/22B and 83630B/L Option 001

7-50 Disassembly and Replacement Procedures

Show 4 more labels / short text on this page
RF Deck Cables · 83620B/22B and 83630B/L · Option 001 · RF Deck
Page 483 of the Agilent Technologies 83640B user manual — 83620B/22B/30B
Page 483 · click to enlarge
Page 483

83620B/22B/30B

Agilent Technologies 83640B — 83620B/22B/30B diagram (page 483)
Fig. 176

Figure 7-42. RF Deck Cable Locations { 83620B/22B/30B Option 006

Agilent 8360 Disassembly and Replacement Procedures

RF Deck Cables · Option 006 · RF Deck

Page 484 of the Agilent Technologies 83640B user manual — 83620B/22B/30B
Page 484 · click to enlarge
Page 484

83620B/22B/30B

Options 001 and 006

Agilent Technologies 83640B — 83620B/22B/30B diagram (page 484)
Fig. 177

Figure 7-43. RF Deck Cable Locations { 83620B/22B/30B Options 001 and 006

7-52 Disassembly and Replacement Procedures

RF Deck Cables · RF Deck

Page 485 of the Agilent Technologies 83640B user manual — 83623B
Page 485 · click to enlarge
Page 485

83623B

Agilent Technologies 83640B — 83623B diagram (page 485)
Fig. 178

Figure 7-44. RF Deck Cable Locations { 83623B

Agilent 8360 Disassembly and Replacement Procedures

RF Deck Cables · No Options · RF Deck

Page 486 of the Agilent Technologies 83640B user manual — 83623B
Page 486 · click to enlarge
Page 486

83623B

Agilent Technologies 83640B — 83623B diagram (page 486)
Fig. 179

Figure 7-45. RF Deck Cable Locations { 83623B Option 001

7-54 Disassembly and Replacement Procedures

RF Deck Cables · Option 001 · RF Deck

Page 487 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 487 · click to enlarge
Page 487
Agilent Technologies 83640B — User Cal Menu diagram (page 487)
Fig. 180 · User Cal Menu

RF Deck Cable Locations { 83623L

Agilent 8360 Disassembly and Replacement Procedures

RF Deck Cables · No Options · RF Deck

Page 488 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 488 · click to enlarge
Page 488
Agilent Technologies 83640B — User Cal Menu diagram (page 488)
Fig. 181 · User Cal Menu

RF Deck Cable Locations { 83623L Option 001

7-56 Disassembly and Replacement Procedures

RF Deck Cables · Option 001 · RF Deck

Page 489 of the Agilent Technologies 83640B user manual — 83624B
Page 489 · click to enlarge
Page 489

83624B

Agilent Technologies 83640B — 83624B diagram (page 489)
Fig. 182

Figure 7-46. RF Deck Cable Locations { 83624B

Agilent 8360 Disassembly and Replacement Procedures

RF Deck Cables · No Options · RF Deck

Page 490 of the Agilent Technologies 83640B user manual — 83624B
Page 490 · click to enlarge
Page 490

83624B

Agilent Technologies 83640B — 83624B diagram (page 490)
Fig. 183

Figure 7-47. RF Deck Cable Locations { 83624B Option 001

7-58 Disassembly and Replacement Procedures

RF Deck Cables · Option 001 · RF Deck

Page 491 of the Agilent Technologies 83640B user manual — 83623B/24B
Page 491 · click to enlarge
Page 491

83623B/24B

Agilent Technologies 83640B — 83623B/24B diagram (page 491)
Fig. 184

Figure 7-48. RF Deck Cable Locations { 83623B/24B Option 006

Agilent 8360 Disassembly and Replacement Procedures

RF Deck Cables · Option 006 · RF Deck

Page 492 of the Agilent Technologies 83640B user manual — 83623B/24B
Page 492 · click to enlarge
Page 492

83623B/24B

Options 001 and 006

Agilent Technologies 83640B — 83623B/24B diagram (page 492)
Fig. 185

Figure 7-49. RF Deck Cable Locations { 83623B/24B Options 001 and 006

7-60 Disassembly and Replacement Procedures

RF Deck Cables · RF Deck

Page 493 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 493 · click to enlarge
Page 493
Agilent Technologies 83640B — User Cal Menu diagram (page 493)
Fig. 186 · User Cal Menu

Figure 7-50. RF Deck Screw Locations

Agilent 8360 Disassembly and Replacement Procedures

RF Deck

Page 494 of the Agilent Technologies 83640B user manual — CA UTION
Page 494 · click to enlarge
Page 494

Motherboard Disassembly and Reassembly

T-15 T orx Screwdriv er

#2 x 4 P ozidriv Screwdriv er

5/16 in. Op en End W renc h

5mm Ball and So c k et Driver

T o disassemble the motherb oard, follo w this pro cedure in the order given.

T o reassemble the motherb oard, follo w this pro cedure in the rev erse order.

CA UTION

Cable routing may a ect instrumen t p erformance. Refer to Figure 7-1 for a

diagram of the instrumen t cable routing.

Refer to the o w c hart in Figure 7-51 to determine the necessary disassembly steps for a given

assembly .

7-62 Disassembly and Replacement Procedures

T ools Required · Description and Procedure · Motherboard

Page 495 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 495 · click to enlarge
Page 495
Agilent Technologies 83640B — User Cal Menu diagram (page 495)
Fig. 187 · User Cal Menu

Figure 7-51. Motherboard Disassembly Flo w chart

Agilent 8360 Disassembly and Replacement Procedures

Motherboard

Page 496 of the Agilent Technologies 83640B user manual — CA UTION
Page 496 · click to enlarge
Page 496
1. Preliminary Steps

CA UTION

Y ou can easily blow the fuse for the 13.5 V supply on the preregulator

assembly if y ou do not disconnect the A C p ow er cord. Also, the danger of

sho c k is increased if the p ow er cord is connected.

1. Disconnect the A C p ow er line cord.
2. Remo v e the screw from eac h of the four feet on the rear panel.
3. Remo v e the four rear feet.
4. Remo v e the four b ottom feet.
5. T o remov e the side straps, remov e the tw o screws on eac h side panel.
6. Remo v e the vin yl trim strip from the top, front edge of the instrumen t. (Place a at

screwdriver in either slot in the trim strip and lift to remov e, see Figure 7-52).

7. Remo v e the instrumen t top, b ottom, and side co v ers.

An RF braid is in eac h of the side c hannels on the top and b ottom of the

instrumen t. When y ou reassemble the instrumen t, b e sure the braids remain

in the side co v er c hannels.

Figure 7-52. Front Edge Trim Strip

7-64 Disassembly and Replacement Procedures

Motherboard

Page 497 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 497 · click to enlarge
Page 497
2. Front P anel Remov al

Refer to Figure 7-53.

1. Complete \1. Preliminary Steps" b efore con tinuing.
2. Remo v e thr e e of the four screws under the trim strip.
3. Remo v e two of the three screws on the low er, front frame edge.
4. Gen tly pry the front panel and display from the frame.

Figure 7-53. Front P anel Screw Locations

Agilent 8360 Disassembly and Replacement Procedures

Motherboard

Page 498 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 498 · click to enlarge
Page 498

Refer to Figure 7-54.

Instrumen ts with Option 004 do not ha v e all the cables sho wn.

5. Cut the tie wraps holding the cables.
6. F rom the A9, A10, and A11 assemblies (or the A8 assembly for instrumen ts with Option

002), disconnect the four coaxial cables (W6, W7, W8, and W9) attached to the BNC

connectors, and the source mo dule in terface cable (W10).

7. Disconnect ribb on cables W2 and W3 from the motherb oard.

Figure 7-54. Front P anel Coaxial and Ribbon Cable Locations

7-66 Disassembly and Replacement Procedures

Motherboard

Page 499 of the Agilent Technologies 83640B user manual — CA UTION
Page 499 · click to enlarge
Page 499
3. J1 RF Output Connector Remov al (Front P anel)

Refer to Figure 7-55.

CA UTION

T o av oid damaging the connector, do not disassemble the RF output

connector assembly . Only unscrew the connector on the end of the cable.

1. Complete \1. Preliminary Steps" and \2. F ront P anel Remo v al" b efore con tinuing.
2. On the low er front frame edge, remov e the tw o RF output screws that hold the RF output

assembly .

3. Remo v e the RF connector with cable W43.
Agilent Technologies 83640B — CA UTION diagram (page 499)
Fig. 188

Figure 7-55. Remo ving RF Output Connector (Front P anel)

Agilent 8360 Disassembly and Replacement Procedures

Motherboard

Page 500 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 500 · click to enlarge
Page 500

When y ou re-attach the connector and cable to the syn thesizer, rst, lo osen

the n ut b et w een cable W43 and the connector. Then rev erse the steps ab ov e.

Finally , torque the n ut b et w een cable W43 and the connector J1 to 10 in-lb

(112 N-cm).

7-68 Disassembly and Replacement Procedures

Motherboard

Page 501 of the Agilent Technologies 83640B user manual — CA UTION
Page 501 · click to enlarge
Page 501
4. J1 RF Output Connector Remov al (Option 004)

Refer to Figure 7-56.

CA UTION

T o av oid damaging the connector, do not disassemble the RF output

connector assembly . Only unscrew the connector on the end of the cable.

1. Complete \1. Preliminary Steps" and \2. F ront P anel Remo v al" b efore con tinuing.
2. Unscrew the RF output connector plate from the rear panel.
3. Remo v e the RF connector with cable W43.

Figure 7-56. Remo ving RF Output Connector (Option 004)

Agilent 8360 Disassembly and Replacement Procedures

Motherboard

Page 502 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 502 · click to enlarge
Page 502
5. Rear P anel Remov al

Refer to Figure 7-57.

1. Complete \1. Preliminary Steps" and \7. J1 RF Output Connector Remo v al" b efore

con tinuing.

2. Remo v e the four screws on the top bac k edge.
3. Remo v e the four screws on the b ottom bac k edge.
4. Slide the rear panel out of the casting.

Figure 7-57. Rear P anel Remo v al

7-70 Disassembly and Replacement Procedures

Motherboard

Page 503 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 503 · click to enlarge
Page 503

Refer to Figure 7-58.

5. Disconnect ribb on cable W31 from the motherb oard.
6. Disconnect the 8-pin line switch connector.
7. Disconnect the fan harness.
8. Cut the tie wraps holding the coaxial cables to the four BNC connectors and the source

mo dule in terface.

9. Disconnect the coaxial cables from the A7, A10, A12, and A14 assemblies and disconnect

the source mo dule in terface cable from the A19 assembly .

Figure 7-58. Rear P anel Cable Locations

Agilent 8360 Disassembly and Replacement Procedures

Motherboard

Page 504 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 504 · click to enlarge
Page 504
6. Major Assemblies Remov al

Refer to Figure 7-59.

1. Complete \1. Preliminary Steps", \2. F ront P anel Remo v al",

\7. J1 RF Output Connector Remo v al", and \5. Rear P anel Remo v al" b efore con tinuing.

2. Remo v e the six screws from the regulator shield.
3. Remo v e the shield.
4. Cut the tie wraps holding the cables attached to the PC assemblies.
5. Disconnect all cables attached to the PC assemblies.
6. Disconnect semi-rigid cable W54 from the A6 assembly .

All cables are lab eled for easy reassembly . F or iden ti cation purp oses, the

extractors on the PC assemblies are color-co ded to indicate their assembly

n um b er.

7. Pull out all PC assemblies.
Agilent Technologies 83640B — User Cal Menu diagram (page 504)
Fig. 189 · User Cal Menu

Figure 7-59. Major Assemblies

7-72 Disassembly and Replacement Procedures

Motherboard

Page 505 of the Agilent Technologies 83640B user manual — CA UTION
Page 505 · click to enlarge
Page 505
7. RF Deck Remov al

Refer to Figure 7-60.

1. Complete \1. Preliminary Steps" and \7. J1 RF Output Connector Remo v al" b efore

con tinuing.

2. Disconnect cable W51 from the attenuator.
3. Remo v e the tw o hex screws from the RF dec k.
Agilent Technologies 83640B — User Cal Menu diagram (page 505)
Fig. 190 · User Cal Menu

Figure 7-60. W51 Location

CA UTION

Av oid damage to the v arious ribb on cables and their connectors attached to

the RF dec k assembly when removing that assembly in the next step. Mov e

all exible coaxial cables out of the w ay of the RF dec k assembly comp onen ts.

4. Carefully pull out the RF dec k and rest it on the side of the instrumen t.
5. Remo v e exible cables W32, W33, W34, and W35 from the motherb oard.
6. Remo v e the hex screws from the RF dec k and set the RF dec k aside.

Agilent 8360 Disassembly and Replacement Procedures

Motherboard

Page 506 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 506 · click to enlarge
Page 506
8. Motherboard Remov al

Refer to Figure 7-61.

1. Complete \1. Preliminary Steps", \2. F ront P anel Remo v al",

\7. J1 RF Output Connector Remo v al", \5. Rear P anel Remo v al", \6. Ma jor Assem blies

Remo v al", and \7. RF Dec k Remo v al" b efore con tinuing.

2. Disconnect the transformer pin-lo c king tab and the reference oscillator pin-lo c king tab.
Agilent Technologies 83640B — User Cal Menu diagram (page 506)
Fig. 191 · User Cal Menu

Figure 7-61. Motherboard Cables

3. Remo v e the 16 screws from the b ottom side of the motherb oard b ottom side Figure 7-62 .
4. Lift o the motherb oard.

7-74 Disassembly and Replacement Procedures

Motherboard

Page 507 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 507 · click to enlarge
Page 507
Agilent Technologies 83640B — User Cal Menu diagram (page 507)
Fig. 192 · User Cal Menu

Figure 7-62. Motherboard Screws

Agilent 8360 Disassembly and Replacement Procedures

Motherboard

Page 508 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 508 · click to enlarge
Page 508
Page 508 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 509 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 509 · click to enlarge
Page 509

T able 8-1.

Adjustments and P erformance T ests Required

After Repair or Replacement of an Assembly

Assembly Adjustmen t Performance T est

A1 F ront P anel Keyb oard None F ull Self{T est

F ront P anel Self{T est

A2 Source Mo dule Interface None F ull Self{T est

A3 F ront P anel Pro cessor None F ull Self{T est

F ront P anel Self{T est

A4 F ractional{N None F ull Self{T est

F requency Switching Time

A5 YO Phase Detector YO Lo op Gain F ull Self{T est

F requency Switching Time

Single Sideband Phase Noise

A6 Sampler Sampler Assem bly F ull Self{T est

F requency Switching Time

Single Sideband Phase Noise

This adjustment or p erformance test is required only for the Agilent 8360 B-Series sw ept signal

generator. Do not p erform this adjustment or test on the 8360 L-Series sw ept CW generator.

The adjustmen ts listed are required only after r ep air of the assembly . The r eplac ement assembly

is adjusted at the factory and is not instrument-dep enden t.

Agilent 8360 P ost{Repair 8-1

Show 7 more labels / short text on this page
P ost{Repair · Spurious Signals · (Non-Harmonics) · Spurious Signals · (Non-Harmonics) · Spurious Signals · (Non-Harmonics)
Page 510 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 510 · click to enlarge
Page 510

T able 8-1.

Adjustments and P erformance T ests Required

After Repair or Replacement of an Assembly (continued)

Assembly Adjustment Performance T est

A7 Reference None F ull Self{T est

F requency Switching Time

Single Sideband Phase Noise

A8 Mo dulation Generator AM/FM D A C O set and Gain

F ull Self{T est

Mo dulation Generator Flatness

A9 Pulse A9 Pulse Bd Jump er

F ull Self{T est

Maximum Leveled P ow er

Low P ow er SRD Bias

A10 ALC Mo dulator O set and Gain

F ull Self{T est

ALC P ow er Level Accuracy

P ow er Accuracy

P ow er Flatness

Maximum Leveled P ow er

A11 FM Driver FM Gain

F ull Self{T est

A12 Multiplier/Filter Driver Amp/Mult Adjustmen ts

F ull Self{T est

Maximum Leveled P ow er

This adjustment or p erformance test is required only for the 8360 B-Series sw ept signal

generator. Do not p erform this adjustment or test on the 8360 L-Series sw ept CW generator.

The adjustments listed are required only after r ep air of the assembly . The r eplac ement assembly

is adjusted at the factory and is not instrument-dep enden t.

8-2 P ost{Repair Agilent 8360

Show 23 more labels / short text on this page
Spurious Signals · (Non-Harmonics) · Spurious Signals · (Line-Related) · Pulse P erformance · AM Accuracy · FM Accuracy · Internal Pulse Accuracy · Mo dulation Meter · Amp/Mult Adjustmen ts · Amp/Filt Adjustmen ts · Pulse P erformance · AM Delay · AM O set · AM Accuracy · External Leveling · AM Accuracy · AM Bandwidth · AM Dynamic Range · FM Accuracy · FM Bandwidth · Maximum FM Deviation · Amp/Filt Adjustmen ts
Page 511 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 511 · click to enlarge
Page 511

T able 8-1.

Adjustments and P erformance T ests Required

After Repair or Replacement of an Assembly (continued)

Assembly Adjustmen t Performance T est

A13 YO Driver YO Driver +10 V Reference

F ull Self{T est

YO Driver Gain and Linearity

Sw ept F requency Accuracy

F requency Switching Time

Single Sideband Phase Noise

A14 Sw eep Generator Sw eep Ramp F ull Self{T est

Sw ept F requency Accuracy

A15 CPU Selftest P atches (L Mo dels only)

F ull Self{T est

A16 P ostregulator None F ull Self{T est

A17 Recti er/Filter None F ull Self{T est

A18 Switching Regulator None F ull Self{T est

A19 Rear P anel Interface None F ull Self{T est

A20 RF Interface None F ull Self{T est

Op eration V eri cation

A21 Alphanumeric Displa y None F ull Self{T est

A22 Motherb oard None F ull Self{T est

Op eration V eri cation

A23 10 MHz Reference

10 MHz Standard F ull Self{T est

Internal TimebaseAging

Single Sideband Phase Noise

This adjustment or p erformance test is required only for the 8360 B-Series sw ept signal

generator. Do not p erform this adjustment or test on the 8360 L-Series sw ept CW generator.

The adjustmen ts listed are required only after r ep air of the assembly . The r eplac ement assembly

is adjusted at the factory and is not instrument-dep enden t.

Agilent 8360 P ost{Repair 8-3

Show 7 more labels / short text on this page
YO Delay (automated) · Maximum FM Deviation · ADC (automated) · GPIB Self{T est · GPIB Self{T est · GPIB Self{T est · Rate
Page 512 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 512 · click to enlarge
Page 512

T able 8-1.

Adjustments and P erformance T ests Required

After Repair or Replacement of an Assembly (continued)

Assembly Adjustment Performance T est

A24 Low Band Mo dulator O set and Gain

F ull Self{T est

ALC P ow er Level Accuracy

P ow er Accuracy

P ow er Flatness

P ow er Flatness

Maximum Leveled P ow er

Single Sideband Phase Noise

Pulse Mo dulation On/O

Pulse Mo dulation Video

A26 YIG Oscillator YO Driver Gain and Linearity

F ull Self{T est

Sw ept F requency Accuracy

F requency Switching Time

Maximum Leveled P ow er

Single Sideband Phase Noise

A28 Amp/Mult Amp/Mult Adjustmen ts

F ull Self{T est

Maximum Leveled P ow er

A29 Amp/Filter Amp/Filter Adjustmen ts F ull Self{T est

Maximum Leveled P ow er

A30 Directional Coupler ALC P ow er Level Accuracy

F ull Self{T est

P ow er Flatness

P ow er Accuracy

Maximum Leveled P ow er

This adjustment or p erformance test is required only for the 8360 B-Series sw ept signal

generator. Do not p erform this adjustment or test on the 8360 L-Series sw ept CW generator.

The adjustments listed are required only after r ep air of the assembly . The r eplac ement assembly

is adjusted at the factory and is not instrument-dep enden t.

8-4 P ost{Repair Agilent 8360

Show 21 more labels / short text on this page
Spurious Signals (Harmonics) · Spurious Signals · (Non-Harmonics) · Ratio · Pulse P erformance · F eedthrough · AM Accuracy · AM Bandwidth · AM Dynamic Range · YO Delay (Automated) · FM Gain · Spurious Signals (Harmonics) · FM Accuracy · FM Bandwidth · Maximum FM Deviation · Amp/Filt Adjustmen ts · Spurious Signals (Harmonics) · FM Bandwidth · Spurious Signals (Harmonics) · Pulse P erformance · FM Bandwidth
Page 513 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 513 · click to enlarge
Page 513

T able 8-1.

Adjustments and P erformance T ests Required

After Repair or Replacement of an Assembly (continued)

Assembly Adjustmen t Performance T est

A30 Bridge Detector ALC P ow er Level Accuracy

F ull Self{T est

P ow er Flatness

P ow er Accuracy

Maximum Leveled P ow er

A31 Step A tten uator ALC P ow er Level Accuracy

F ull Self{T est

Step A tten uator Flatness

Maximum Leveled P ow er

A tten uator V eri cation

(only if data is required)

A32 Doubler P ow er Flatness F ull Self{T est

P ow er Accuracy

P ow er Flatness

Maximum Leveled P ow er

A33 Ampli er/Switch P ow er Flatness F ull Self{T est

Maximum Leveled P ow er

Pulse Mo dulation On/O

A33 Ampli er P ow er Flatness F ull Self{T est

Maximum Leveled P ow er

Pulse Mo dulation On/O

A34 Rear P anel BNC None F ull Self{T est

A38 Dual Mo dulator Amp/Filter Adjustments

Maximum leveled P ow er

A38 Mo d Doubler Amp Mo dulator O set and Gain

ALC P ow er Level Accuracy

Pulse Mo dulation On/O

P ow er Flatness

Pulse Mo dulation Video

This adjustment or p erformance test is required only for the 8360 B-Series sw ept signal

generator. Do not p erform this adjustment or test on the 8360 L-Series sw ept CW generator.

The adjustmen ts listed are required only after r ep air of the assembly . The r eplac ement assembly

is adjusted at the factory and is not instrument-dep enden t.

The entire ampli er/ lter adjustment pro cedure do es not need to b e p erformed. P erform only

the squegging clamp p ortions of this adjustment.

Agilent 8360 P ost{Repair 8-5

Show 15 more labels / short text on this page
(automated) · Spurious Signals (Harmonics) · Spurious Signals (Harmonics) · Ratio · Spurious Signals (Harmonics) · Ratio · Spurious Signals (Harmonics) · Ratio · AM Accuracy · Pulse P erformance · AM O set · F eedthrough · AM Accuracy · AM Bandwidth · AM Dynamic Range
Page 514 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 514 · click to enlarge
Page 514

T able 8-1.

Adjustments and P erformance T ests Required

After Repair or Replacement of an Assembly (continued)

Assembly Adjustment Performance T est

A39 Directional Coupler None F ull Self{T est

Maximum Leveled P ow er

A41 Dio de Switch None F ull Self{T est

Maximum Leveled P ow er

A42 Pulse Mo d Switched Filter Mo dulator O set and Gain

Maximum Leveled P ow er

Pulse Mo dulation On/O

Pulse Mo dulation Video

B1 F an None Spurious Signals

CR1 Detector ALC P ow er Level Accuracy

F ull Self{T est

P ow er Flatness

P ow er Accuracy

Maximum Leveled P ow er

FL1 Line Mo dule None Spurious Signals

FL2 8 GHz LPF Sampler Assem bly F ull Self{T est

J1 RF Output Connector P ow er Flatness (automated)

F ull Self{T est

Step A tten uator (automated)

Maximum Leveled P ow er

RPG1 Rotary Pulse Generator None F ull Self{T est

F ront P anel Self{T est

T1 T oroid P ow er T ransformer None F ull Self{T est

This adjustment or p erformance test is required only for the 8360 B-Series sw ept signal

generator. Do not p erform this adjustment or test on the 8360 L-Series sw ept CW generator.

The adjustments listed are required only after r ep air of the assembly . The r eplac ement assembly

is adjusted at the factory and is not instrument-dep enden t.

8-6 P ost{Repair Agilent 8360

Show 8 more labels / short text on this page
Spurious Signals (Harmonics) · Ratio · Pulse P erformance · F eedthrough · AM Bandwidth · AM Dynamic Range · (Non-Harmonics) · (Line-Related)
Page 515 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 515 · click to enlarge
Page 515

Replaceable parts include ma jor assemblies and all c hassis hardw are. T able 9-1 lists reference

designations and abbreviations used in this c hapter.

F or information on removing and replacing assemblies, see Chapter 7, \Disassembly and

Replacemen t Pro cedures."

Many ma jor assemblies are co v ered b y the mo dule{exc hange program. Under the terms of

the program, factory{repaired and tested assemblies are av ailable on a trade{in basis (a

defectiv e assembly must b e returned for credit). Exc hange assemblies meet all new assembly

sp eci cations, but are less exp ensiv e. Figure 9-1 illustrates the mo dule exc hange pro cedure.

In order to prev en t damage when returning pro ducts to Agilent T ec hnologies, use original or

comparable pac k aging.

Order spare assembly sto c k using the new assembly part n um b er.

If y ou ha v e an y questions, con tact y our Agilent T ec hnologies customer engineer for the latest

information ab out this program.

The follo wing information is given in the tables in this c hapter:

Agilent part n um b er.

P art quan tity for that gure. There may b e more of that part in other gures.

P art description (see T able 9-1 for abbreviations).

Agilent 8360 Replaceable P arts 9-1

Show 4 more labels / short text on this page
Replaceable P arts · Introduction · Module{Exchange Program · Replacement Information
Page 516 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 516 · click to enlarge
Page 516

F or an y listed part, request the Agilent part n um b er and quan tity required from the nearest

Agilent oce.

Ho w to Order P arts . . . F ast!

Agilent parts sp ecialists ha v e direct on{line access to the replaceable parts listed in this

man ual. F our{day delivery is standard; there is a c harge for hotline (one{day) delivery .

In the United States, call the follo wing toll{free n um b er: (800) 227{8164

Monday through F riday , 6 am to 5 pm (Paci c Standard Time)

Outside the United States, con tact y our nearest Agilent oce.

9-2 Replaceable P arts Agilent 8360

Ordering Information

Page 517 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 517 · click to enlarge
Page 517

T able 9-1. Reference Designations and Abbreviations

A Amp erage; Assem bly

A T T ermination

LPF Low P ass Filter

A Y Assem bly

M Male; Maximum; Meter; Mil; Milli

B F an; Motor

P AN{HD P an Head

CMPNT Comp onen t

RP Rear P anel

RPG Rotary Pulse Generator

SKT So ck et

SRD Step Recovery Dio de

T T ransformer Thickness; T ap er;

FP F ront P anel

HD Hand; Hard; Heavy Duty; Head

TR{HD T russ Head

HEX Hexadecimal; Hexagon; Hexagonal

UL Underwriters Lab oratories

YIG Yttrium Iron Garnate

J Electrical Connector (Stationary

Agilent 8360 Replaceable P arts 9-3

Show 37 more labels / short text on this page
LG Long; Length · LK Lo ck · AX Axial Lead · MTG Mounting · BI Bipin · NTD Non-Time Delay · BT Battery · NYL Nylon · CHAM Chamfer · CONN Connector · CR Detector · CU Copp er · SN{PL Tin Plated · D Diameter · DEG Degree · STD Standard · F F use · STL Steel · FE F errule · SW Switch · FL Filter · T o oth · FLH Flat Head · TD Time Delay · FL TG Floating · THD Thread; Threaded · FR F ront · THK Thick · TPG T apping · ID Inside Diameter · IN Inch · W Cable; Wire · INTL Internal; International · WD Width · P osition); Jack · YO YIG Oscillator · LBL Lab el
Page 518 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 518 · click to enlarge
Page 518
Agilent Technologies 83640B — User Cal Menu diagram (page 518)
Fig. 193 · User Cal Menu

Figure 9-1. Module Exchange Program

9-4 Replaceable P arts Agilent 8360

Page 519 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 519 · click to enlarge
Page 519

Major Assemblies{T op View

Agilent Technologies 83640B — User Cal Menu diagram (page 519)
Fig. 194 · User Cal Menu

Figure 9-2. Major Assemblies (1 of 5)

Agilent 8360 Replaceable P arts 9-5

Page 520 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 520 · click to enlarge
Page 520
Agilent Technologies 83640B — User Cal Menu diagram (page 520)
Fig. 195 · User Cal Menu

Figure 9-2. Major Assemblies (2 of 5)

9-6 Replaceable P arts Agilent 8360

Major Assemblies{RF Deck

Page 521 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 521 · click to enlarge
Page 521
Agilent Technologies 83640B — User Cal Menu diagram (page 521)
Fig. 196 · User Cal Menu

Figure 9-2. Major Assemblies (3 of 5)

Agilent 8360 Replaceable P arts 9-7

Major Assemblies{RF Deck

Page 522 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 522 · click to enlarge
Page 522
Agilent Technologies 83640B — User Cal Menu diagram (page 522)
Fig. 197 · User Cal Menu

Figure 9-2. Major Assemblies (4 of 5)

9-8 Replaceable P arts Agilent 8360

Major Assemblies{RF Deck

Page 523 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 523 · click to enlarge
Page 523
Agilent Technologies 83640B — User Cal Menu diagram (page 523)
Fig. 198 · User Cal Menu

Figure 9-2. Major Assemblies (5 of 5)

Agilent 8360 Replaceable P arts 9-9

Major Assemblies{RF Deck

Page 524 of the Agilent Technologies 83640B user manual — A10W1 1258-0124 1 JUMPER{AM INPUT IMPED ANCE
Page 524 · click to enlarge
Page 524

T able 9-2. Major Assemblies

Ref.

Desig.

A1 08360-60001 1 FRONT P ANEL KEYBO ARD ASSEMBL Y

A2 08360-60002 1 SOUR CE MODULE INTERF A CE ASSEMBL Y

A2S1 3101-2995 1 PO WER ON/OFF SWITCH

A3 06360-60003 1 FRONT P ANEL PR OCESSOR ASSEMBL Y

A4 08360-60295 1 FRA CTIONAL{N ASSEMBL Y

A5 08360-60209 1 YO LOOP ASSEMBL Y

A5 08360-69209 YO LOOP ASSEMBL Y (Rebuilt{Exchange)

A6 08360-60214 1 SAMPLER ASSEMBL Y

A6 08360-69214 SAMPLER ASSEMBL Y (Rebuilt{Exchange)

A7 08360-60008 1 REFERENCE ASSEMBL Y

A7 08360-69008 REFERENCE ASSEMBL Y (Rebuilt{Exchange)

A8 08360-60166 1 MODULA TION GENERA TION ASSEMBL Y (Option 002)

(Agilent 8360 B-series only)

A8 08360-69166 MODULA TION GENERA TION ASSEMBL Y (Rebuilt{Exchange)

A8W1 1258-0124 1 JUMPER{AM INPUT IMPED ANCE (8360 B-series only)

A8W2 1258-0124 1 JUMPER{FM INPUT IMPED ANCE (8360 B-series only)

A9 08360-60294 1 PULSE ASSEMBL Y REPLA CEMENT KIT (Contains 08360-60292 and

A10 08360-60222 1 ALC ASSEMBL Y (Agilent 83620B/22B/24B)

A10 08360-69222 ALC ASSEMBL Y (Rebuilt{Exchange)

A10 08360-60206 1 ALC ASSEMBL Y (Agilent 83640B/L and Agilent 83650B/L/23L/30L)

A10 08360-69206 ALC ASSEMBL Y (Rebuilt{Exchange)

A10W1 1258-0124 1 JUMPER{AM INPUT IMPED ANCE

A11 08360-60011 1 FM DRIVER ASSEMBL Y (8360 B-series only)

A11W1 8159-0005 1 ZER O OHM RESISTOR{FM INPUT IMPED ANCE (8360 B-series only)

9-10 Replaceable P arts Agilent 8360

Show 5 more labels / short text on this page
P art · Qty Description · Num b er · (8360 B-series only) · Installation Note)
Page 525 of the Agilent Technologies 83640B user manual — A18 08360-60198 1 SWITCHING REGULA TOR ASSEMBL Y
Page 525 · click to enlarge
Page 525

T able 9-2. Major Assemblies (continued)

Ref.

Desig.

A12 08360-60212 1 MUL TIPLIER/FIL TER DRIVER ASSEMBL Y

A12 08360-69212 MUL TIPLIER/FIL TER DRIVER ASSEMBL Y (Rebuilt{Exchange)

A13 08360-60207 1 YO DRIVER ASSEMBL Y

A13 08360-69207 YO DRIVER ASSEMBL Y (Rebuilt{Exc hange)

A14 08360-60014 1 SWEEP GENERA TOR ASSEMBL Y

A14 08360-69014 SWEEP GENERA TOR ASSEMBL Y (Rebuilt{Exchange)

A15 08360-60264 1 CPU ASSEMBL Y

A15 08360-69264 CPU ASSEMBL Y (Rebuilt{Exchange)

08360-60201 1 FIRMW ARE UPGRADE KIT

A15BT1 1420-0394 1 BA TTER Y (Used with CPU Assem bly 08360-60264)

A16 08360-60016 1 POSTREGULA TOR ASSEMBL Y

A16 08360-69016 POSTREGULA TOR ASSEMBL Y (Rebuilt-Exchange)

0400-0018 1 GR OMMET-CHAN .052 TH (P art of A16)

A16F1 2110-0047 1 FUSE (INCH) 1A 125V NTD BI

A16F2 2110-0618 3 FUSE (INCH) 5A 125V NTD BI

A16F3 2110-0713 2 FUSE{SUBMINIA TURE 10A 125V NTD AX

A16F4 2110-0713 FUSE{SUBMINIA TURE 10A 125V NTD AX

A16F5 2110-0618 FUSE (INCH) 5A 125V NTD BI

A16F6 2110-0618 FUSE (INCH) 5A 125V NTD BI

A16F7 2110-0425 2 FUSE (INCH) 2A 125V NTD BI

A16F8 2110-0425 FUSE (INCH) 2A 125V NTD BI

A17 08360-60017 1 RECTIFIER/FIL TER ASSEMBL Y

A17 08360-69017 RECTIFIER/FIL TER ASSEMBL Y (Rebuilt{Exchange)

A17F1 2110-0303 1 FUSE (INCH) 2A 250V TD FE UL

A18 08360-60198 1 SWITCHING REGULA TOR ASSEMBL Y

A18 08360-69198 SWITCHING REGULA TOR ASSEMBL Y (Rebuilt{Exchange)

A18F1 2110-0201 1 FUSE (INCH) 0.25A 250V TD FE UL

Agilent 8360 Replaceable P arts 9-11

P art · Qty Description · Num b er

Page 526 of the Agilent Technologies 83640B user manual — A29 5086-7637 1 AMPLIFIER/FIL TER (83623L)
Page 526 · click to enlarge
Page 526

T able 9-2. Major Assemblies (continued)

Ref.

Desig.

A19 08360-60019 1 REAR P ANEL INTERF A CE ASSEMBL Y

A20 08360-60266 1 RF INTERF A CE ASSEMBL Y

A20F1 2110-0425 1 FUSE (INCH) 2A 125V NTD BI

A21 2090-0357 1 ALPHANUMERIC DISPLA Y ASSEMBL Y

A22 08360-60196 1 MOTHERBOARD ASSEMBL Y

A23 10811E 1 10 MHZ REFERENCE ST AND ARD (Includes W30)

A23 1813-1258 1 OCX O 10 MHz

A24 5086-7463 1 LO WBAND ASSEMBL Y

A24 5086-6463 LO WBAND ASSEMBL Y (Rebuilt{Exc hange)

A25 NOT USED

A26 5086-7515 1 YIG OSCILLA TOR (2.0{7.5 GHz)

A26 5086-6515 YIG OSCILLA TOR (2.0{7.5 GHz) (Rebuilt{Exc hange)

A27 NOT USED

A28 5086-7622 1 AMPLIFIER/MUL TIPLIER

A28 5086-6622 AMPLIFIER/MUL TIPLIER (Rebuilt{Exchange)

A29 5086-7616 1 FIL TER (Agilent 83620B/22B/23B/24B)

A29 5086-6616 FIL TER (Rebuilt{Exchange)

A29 5086-7637 1 AMPLIFIER/FIL TER (83623L)

A29 5086-7623 1 AMPLIFIER/FIL TER

(Agilent 83630B/L, 83640B/L, and 83650B /L)

A29 5086-6623 AMPLIFIER/FIL TER (Rebuilt{Exchange)

A29 5086-7636 1 AMPLIFIER/FIL TER (83630B /40B/50B Option 006)

A29 5086-6636 AMPLIFIER/FIL TER (Rebuilt{Exchange)

A30 5086-7645 1 BRIDGE DETECTOR (83620B/22B and 83630B /L)

A30 0955-0148 1 W A VE DIRECTIONAL COUPLER 20 GHZ MAX

A30 5086-7644 1 BRIDGE DETECTOR (83640B/L and 83650B/L)

A30 5086-7656 1 BRIDGE DETECTOR (83620B/22B/30B Option 006)

9-12 Replaceable P arts Agilent 8360

Show 4 more labels / short text on this page
P art · Qty Description · Num b er · (83623B/L and 83624B)
Page 527 of the Agilent Technologies 83640B user manual — A39 0955-0098 1 COUPLER (2.0 - 8.4 GHZ)
Page 527 · click to enlarge
Page 527

T able 9-2. Major Assemblies (continued)

Ref.

Desig.

A31 33326-60003 STEP A TTENUA TOR

(83620B/22B/24B, 83623B/L, and 83630B/L)

A31 33326-60002 1 STEP A TTENUA TOR (83640B/L)

A31 33326-60005 1 STEP A TTENUA TOR (83650B/L)

A32 5086-7607 1 DOUBLER (83640B/L)

A32 5086-6607 DOUBLER (Rebuilt{Exchange)

A32 5086-7604 1 DOUBLER (83650B/L

A32 5086-6604 DOUBLER (Rebuilt{Exchange)

A33 5086-7465 1 AMPLIFIER/SWITCH (83623B)

A33 5086-6465 1 AMPLIFIER/SWITCH (Rebuilt{Exchange)

A33 5086-7510 1 AMPLIFIER (2{20 GHz) (83624B)

A33 5086-7610 1 AMPLIFIER (2{20 GHz) (Rebuilt{Exchange)

A34 08360-60061 1 RP FLOA TING BNC ASSEMBL Y (Option 004)

A34W1 08360-60099 1 EXTERNAL LEVELING A34J3/A10J4 (Option 004)

A34W2 08360-60100 1 AM INPUT A34J1/A10J8 (Option 004)

A34W3 08360-60101 1 FM INPUT A34J2/A11J1 (Option 004)

A34 08360-60186 1 RP FLOA TING BNC ASSEMBL Y (Option 002 and 004)

A34W1 08360-60099 1 EXTERNAL LEVELING A34J3/A10J4 (Option 002 and 004)

A34W2 08360-60178 1 AM INPUT A34J1/A8J7 (Option 002 and 004)

A34W3 08360-60179 1 FM INPUT A34J2/A8J10 (Option 002 and 004)

A35 NOT USED

A36 NOT USED

A38 5086-7602 1 DUAL MODULA TOR (B Mo del)

A38 5086-7634 1 AM MODULA TOR (8360 B-series Option 006 and 83623L)

A38 5087-7017 1 Mo d Doubler Amp (83630L/40L/50L)

A39 0955-0098 1 COUPLER (2.0 - 8.4 GHZ)

A40 11742A 1 DC BLOCKING CAP A CITOR (83623L )

A41 5086-7327 1 DIODE SWITCH (All mo dels except 83624B)

Agilent 8360 Replaceable P arts 9-13

P art · Qty Description · Num b er

Page 528 of the Agilent Technologies 83640B user manual — BLOCKING CAP A CITOR
Page 528 · click to enlarge
Page 528

T able 9-2. Major Assemblies (continued)

Ref.

Desig.

A42 5086-7628 1 PULSE MODULA TION SWITCHED FIL TER (83620B/22B

A42 5086-7629 1 PULSE MODULA TION SWITCHED FIL TER (83623B/24B

A42 5086-7630 1 PULSE MODULA TION SWITCHED FIL TER (83630B Option 006)

A42 5086-7631 1 PULSE MODULA TION SWITCHED FIL TER (83640B/50B

A T2 1810-0118 1 TERMINA TION SMA MALE 50

A T3 0955-0216 1 A TTENUA TOR 20 DB

B1 08360-80026 1 F AN

CR1 08360-60147 1 DETECTOR REPLA CEMENT KIT (83623B/L and 83624B)

F1 2110-0227 1 FUSE (INCH) 5A 250V NTD FE UL (See T able 9-7 )

F1 2110-0003 1 FUSE (INCH) 3A 230V NTD FE UL (See T able 9-7 )

FL1 9135-0242 1 LINE MODULE { FIL TERED

FL2 0955-0485 1 8 GHZ LO W P ASS FIL TER

J1 08673-60040 1 RF OUTPUT CONNECTOR

(83620B/22B/24B, 83623B /L, and 83630B/L)

J1 5063-1700 1 RF OUTPUT CONNECTOR (83640B/L and 83650B/L) { F or serial

pre xes 4040A, also order W43

T1 9100-4781 1 TOR OID PO WER TRANSFORMER

1250-1158 1 AD APTER SMA(F)|SMA(F) (83623L) USED WITH A40 DC

BLOCKING CAP A CITOR

9-14 Replaceable P arts Agilent 8360

Show 6 more labels / short text on this page
P art · Qty Description · Num b er · Option 006 ) · Option 006 ) · Option 006 )
Page 529 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 529 · click to enlarge
Page 529
Agilent Technologies 83640B — User Cal Menu diagram (page 529)
Fig. 199 · User Cal Menu

Figure 9-3. Cables (1 of 19)

Agilent 8360 Replaceable P arts 9-15

Cables{Front P anel

Page 530 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 530 · click to enlarge
Page 530
Agilent Technologies 83640B — User Cal Menu diagram (page 530)
Fig. 200 · User Cal Menu

Figure 9-3. Cables (2 of 19)

9-16 Replaceable P arts Agilent 8360

Cables{T op View

Page 531 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 531 · click to enlarge
Page 531

8360 B-Series Option 002

Agilent Technologies 83640B — User Cal Menu diagram (page 531)
Fig. 201 · User Cal Menu

Figure 9-3. Cables (3 of 19)

Agilent 8360 Replaceable P arts 9-17

Cables{T op View

Page 532 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 532 · click to enlarge
Page 532

Cables{83640B/L and 83650B/L RF Deck

Agilent Technologies 83640B — User Cal Menu diagram (page 532)
Fig. 202 · User Cal Menu

Figure 9-3. Cables (4 of 19)

9-18 Replaceable P arts Agilent 8360

No Options

Page 533 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 533 · click to enlarge
Page 533

Cables{83640B/L and 83650B/L RF Deck

Agilent Technologies 83640B — User Cal Menu diagram (page 533)
Fig. 203 · User Cal Menu

Figure 9-3. Cables (5 of 19)

Agilent 8360 Replaceable P arts 9-19

Option 001

Page 534 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 534 · click to enlarge
Page 534
Agilent Technologies 83640B — User Cal Menu diagram (page 534)
Fig. 204 · User Cal Menu

Figure 9-3. Cables (6 of 19)

9-20 Replaceable P arts Agilent 8360

Cables{83640B/50B RF Deck · Option 006

Page 535 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 535 · click to enlarge
Page 535

Options 001 and 006

Agilent Technologies 83640B — User Cal Menu diagram (page 535)
Fig. 205 · User Cal Menu

Figure 9-3. Cables (7 of 19)

Agilent 8360 Replaceable P arts 9-21

Cables{83640B/50B RF Deck

Page 536 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 536 · click to enlarge
Page 536

Cables{83620B/22B and 83630B/L RF Deck

Agilent Technologies 83640B — User Cal Menu diagram (page 536)
Fig. 206 · User Cal Menu

Figure 9-3. Cables (8 of 19)

9-22 Replaceable P arts Agilent 8360

No Options

Page 537 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 537 · click to enlarge
Page 537

Cables{83620B/22B and 83630B/L RF Deck

Agilent Technologies 83640B — User Cal Menu diagram (page 537)
Fig. 207 · User Cal Menu

Figure 9-3. Cables (9 of 19)

Agilent 8360 Replaceable P arts 9-23

Option 001

Page 538 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 538 · click to enlarge
Page 538
Agilent Technologies 83640B — User Cal Menu diagram (page 538)
Fig. 208 · User Cal Menu

Figure 9-3. Cables (10 of 19)

9-24 Replaceable P arts Agilent 8360

Cables{83620B/22B/30B RF Deck · Option 006

Page 539 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 539 · click to enlarge
Page 539

Options 001 and 006

Agilent Technologies 83640B — User Cal Menu diagram (page 539)
Fig. 209 · User Cal Menu

Figure 9-3. Cables (11 of 19)

Agilent 8360 Replaceable P arts 9-25

Cables{83620B/22B/30B RF Deck

Page 540 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 540 · click to enlarge
Page 540
Agilent Technologies 83640B — User Cal Menu diagram (page 540)
Fig. 210 · User Cal Menu

Figure 9-3. Cables (12 of 19)

9-26 Replaceable P arts Agilent 8360

Cables{83623B RF Deck · No Options

Page 541 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 541 · click to enlarge
Page 541
Agilent Technologies 83640B — User Cal Menu diagram (page 541)
Fig. 211 · User Cal Menu

Figure 9-3. Cables (13 of 19)

Agilent 8360 Replaceable P arts 9-27

Cables{83623B RF Deck · Option 001

Page 542 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 542 · click to enlarge
Page 542
Agilent Technologies 83640B — User Cal Menu diagram (page 542)
Fig. 212 · User Cal Menu

Figure 9-3. Cables (14 of 19)

9-28 Replaceable P arts Agilent 8360

Cables{83624B RF Deck · No Options

Page 543 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 543 · click to enlarge
Page 543
Agilent Technologies 83640B — User Cal Menu diagram (page 543)
Fig. 213 · User Cal Menu

Figure 9-3. Cables (15 of 19)

Agilent 8360 Replaceable P arts 9-29

Cables{83624B RF Deck · Option 001

Page 544 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 544 · click to enlarge
Page 544
Agilent Technologies 83640B — User Cal Menu diagram (page 544)
Fig. 214 · User Cal Menu

Figure 9-3. Cables (16 of 19)

9-30 Replaceable P arts Agilent 8360

Cables{83623B/24B RF Deck · Option 006

Page 545 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 545 · click to enlarge
Page 545

Options 001 and 006

Agilent Technologies 83640B — User Cal Menu diagram (page 545)
Fig. 215 · User Cal Menu

Figure 9-3. Cables (17 of 19)

Agilent 8360 Replaceable P arts 9-31

Cables{83623B/24B RF Deck

Page 546 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 546 · click to enlarge
Page 546

Figure 9-3. Cables (18 of 19)

9-32 Replaceable P arts Agilent 8360

Cables{83623L RF Deck · No Options

Page 547 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 547 · click to enlarge
Page 547
Agilent Technologies 83640B — User Cal Menu diagram (page 547)
Fig. 216 · User Cal Menu

Figure 9-3. Cables (19 of 19)

Agilent 8360 Replaceable P arts 9-33

Cables{83623L RF Deck · Option 001

Page 548 of the Agilent Technologies 83640B user manual — W20 08360-60082 1 100 MHZ REF{A7J7/A24J2
Page 548 · click to enlarge
Page 548

T able 9-3. Cables

Ref.

Desig.

W1 08360-60062 1 DISPLA Y PO WER A3J1/A21J1

W2 08360-60056 1 SMI/MOTHERBO ARD{A2J1/A22J1

W3 08360-60057 1 F.P . PR OCESSOR/MOTHERBO ARD A3J2/A22J2

W4 08360-60055 1 F.P . PR OCESSOR/DISPLA Y A3J3/A21J2

W5 08360-60054 1 KEYBOARD/F.P . PR OCESSOR A1J1/A3J4

W6 08360-60070 1 EXTERNAL PULSE J2/A9J4 (8360 B-Series)

W6 08360-60094 1 EXTERNAL PULSE J8/A9J4 (8360 B-Series Option 004)

W6 08360-60172 1 EXTERNAL PULSE J2/A8J3 (8360 B-Series Option 002)

W6 08360-60175 1 EXTERNAL PULSE J8/A8J3 (8360 B-Series Option 002 and 004)

W7 08360-60068 1 AM INPUT J3/A10J8 (8360 B-Series)

W7 08360-60173 1 AM INPUT J3/A8J7 (8360 B-Series Option 002)

W8 08360-60069 1 FM INPUT J4/A11J1 (8360 B-Series)

W8 08360-60174 1 FM INPUT J4/A8J10 (8360 B-Series Option 002)

W9 08360-60051 1 EXTERNAL LEVELING J5/A10J4

W10 08360-60092 1 FP SMI LEVELING{A2J2/A10J2

W11 08360-60301 1 10 MHZ REF{A7J5/A4J1

W12 08360-60302 1 30{60 MHZ{A4J3/A5J2

W13 08360-60076 1 SAMPLER IF{A6J2/A5J1

W14 08360-60083 1 HIGH FREQ PLL{A5J3/A26J2

W15 08360-60072 1 40 MHZ REF{A7J6/A6J1

W15 08360-60182 1 40 MHz REF{A7J6/A8J1 (8360 B-Series Option 002)

W16 08360-60095 1 EXTERNAL STD{J6/A7J1

W17 08360-60071 1 O VEN STD{A23J1/A7J2

W18 08360-60073 1 1 MHZ{A7J4/A14J1

W19 08360-60303 1 10 MHZ REAR P ANEL {A4J2/J7

W20 08360-60082 1 100 MHZ REF{A7J7/A24J2

9-34 Replaceable P arts Agilent 8360

P art · Qty Description · Num b er

Page 549 of the Agilent Technologies 83640B user manual — W39 08360-60046 1 RF INTERF A CE/LO WBAND{A20J4/A24J1
Page 549 · click to enlarge
Page 549

T able 9-3. Cables (continued)

Ref.

Desig.

W21 NOT USED

W22 08360-60228 1 PRE-PULSE LEVELING DET{A38J13/A10J6

W22 08360-60245 1 PRE-PULSE LEVELING DET{A42J4/A10J6

(8360 B-Series Option 006)

W23 08360-60093 1 RP SMI LEVELING{A19J2/A10J3

W24 08360-60107 1 LO WBAND DETECTOR{A24J3/A10J5

W25 08360-60106 1 HI BAND DETECTOR{CR1/A10J7

W26 08360-60079 1 YO FM SIGNAL A11J2/A26J3 (8360 B-Series only)

W27 08360-60097 1 V/GHZ{A12J1/J10

W28 08360-60098 1 TRIGGER OUTPUT{A14J2/J9

W29 08360-60031 1 PREREG IN/PREREG OUT{A18J1/A17J1

W30 08360-60063 1 MOTHERBO ARD/10MHZ STD{A22J7/A23J2

W31 08360-60037 1 REAR P ANEL/MOTHERBO ARD{A19J1/A22J8

W32 08360-60226 1 PULSE MOD HI{A22J3/A38J3

W32 08360-60247 1 SLO W PULSE{A22J3/A42J3 (8360 B- Series Option 006)

W33 08360-60088 1 PULSE MOD LO{A22J4/A24J6 (not used on 83623L/30L/40L/50L)

W34 08360-60087 1 ALC MOD LO W{A22J5/A24J7

W35 08360-60225 1 ALC MOD HI{A22J6/A38J2

W36 08360-60230 1 RF INTERF A CE/AMP SWITCH{A20J14/A33J1 (83623B)

W36 08360-60230 1 RF INTERF A CE/AMPLIFIER{A20J14/A33J1 (83624B)

W36 08360-60191 1 RF INTERF A CE/DOUBLER{A20J5/A32J1 (83640B/L and 83650B/L)

W37 08360-60189 1 RF INTERF A CE/AMP MUL T{A20J9/A28J1

W38 08360-60227 1 RF INTERF A CE/DUAL MOD{A20J1/A38J1

W38 08360-60227 1 RF INTERF A CE/AM MOD{A20J1/A38J1

(8360 B-Series Option 006)

W39 08360-60046 1 RF INTERF A CE/LO WBAND{A20J4/A24J1

W40 08360-60045 1 RF INTERF A CE/A TTEN A20J6/A31J1 (Option 001)

Agilent 8360 Replaceable P arts 9-35

P art · Qty Description · Num b er

Page 550 of the Agilent Technologies 83640B user manual — (83623L)
Page 550 · click to enlarge
Page 550

T able 9-3. Cables (continued)

Ref.

Desig.

W41 08360-60190 1 RF INTERF A CE/AMP FIL TER{A20J8/A29J1

W42 08360-60040 1 RF INTERF A CE/YO{A20J10/A26J1

W43 08360-20072 1 DIR COUPLER/RF OUT{A30J2/J1 (83620B/22B/24B, 83623B/L, and

83630B/L)

W43 08360-20072 1 A TTEN/RF OUT{A31J2/J1 (83620B/22B/24B, 83623B/L, and

W43 08360-20073 1 DIR COUPLER/RF OUT{A30J2/J1 (83620B/22B/24B, 83623B/L, and

W43 08360-20073 1 A TTEN/RF OUT{A31J2/J1 (83620B/22B/24B, 83623B/L, and

W43 08360-20097 1 DIR COUPLER/RF OUT{A30J2/J1 (83640B/L and 83650B/L)

W43 08360-20097 1 A TTEN/RF OUT{A31J2/J1

(83640B/L and 83650B/L Option 001)

W43 08360-20098 1 DIR COUPLER/RF OUT{A30J2/J1 (83640B/L and 83650B/L Option

W43 08360-20098 1 A TTEN/RF OUT{A31J2/J1 (83640B/L and 83650B/L Option 014)

W44 08360-20191 1 BRIDGE DET/A TTEN A30J5/A31J1 (83620B /22B/30B Options 001

W44 08360-20071 1 DIR COUPLER/A TTEN A30J2/A31J1 (83623B /24B Options 001 and

W44 08360-20193 1 BRIDGE DET/A TTEN A30J5/A31J1 (83640B /L and 83650B /L Option

W45 08360-20185 1 AMP FIL TER/BRIDGE DET{A29J3/A30J4 (83620B/22B and

83630B/L)

W45 08360-20190 1 AMP FIL TER/BRIDGE DET{A29J3/A30J4 (83620B/22B and

W45 08360-20189 1 DC BLOCKING CAP A CITOR/DIR COUPLER{A40J2/A30J1

(83623L)

W45 08360-20186 1 DC BLOCKING CAP A CITOR/DIR COUPLER{A40J2/A30J1

9-36 Replaceable P arts Agilent 8360

Show 12 more labels / short text on this page
P art · Qty Description · Num b er · 83630B/L Option 001) · 83630B/L Option 004) · 83630B/L Option 014) · 004) · and 006) · 006) · 001) · 83630B/L Option 001) · (83623L Option 001)
Page 551 of the Agilent Technologies 83640B user manual — 83624B)
Page 551 · click to enlarge
Page 551

T able 9-3. Cables (continued)

Ref.

Desig.

W46 NOT USED

W47 08360-20075 1 LO WBAND/AMP SWITCH{A24J4/A33J3 (83623B )

W47 08360-20163 1 LO WBAND/AMP FIL TER{A24J4/A29J2 (83620B/22B, 83623L, and

83630B/L)

W47 08360-20139 1 LO WBAND/DOUBLER{A24J4/A32J3 (83640B/L and 83650B/L)

W47 08360-20246 1 LO WBAND/PLS MOD SW FIL T{A24J4/A42J7 (83620B/23B/30B

W48 08360-20235 1 AMP MUL T/A UX OUTPUT{A28J4/J14

W48 08360-20236 1 AMP MUL T/A UX OUTPUT{A28J4/J14 (Option 004)

W49 08360-20171 1 DIODE SWITCH/AMP MUL T{A41J3/A28J5 (All mo dels except

83624B)

W49 08360-20230 1 COUPLER/AMP MUL T{A39J2/A28J5 (83624B)

W50 08360-20169 1 DIODE SWITCH/LO WBAND{A41J2/A24J5 (All mo dels except

83624B)

W51 08360-20173 1 A TTENUA TOR/LPF{A T3J2/FL2J1

W51 08360-20172 1 A TTENUA TOR/LPF{A T3J2/FL2J1 (Option 004)

W52 08360-20147 1 YO/DIR COUPLER{A26J4/A39J1

W53 NOT USED

W54 08360-20087 1 LPF/SAMPLER{FL2J2/A6J3

W55 1250-1159 1 DIR COUPLER/DIODE SWITCH{A39J2/A41J1 (All mo dels except

83624B)

Agilent 8360 Replaceable P arts 9-37

Show 4 more labels / short text on this page
P art · Qty Description · Num b er · Option 006)
Page 552 of the Agilent Technologies 83640B user manual — W60 NOT USED
Page 552 · click to enlarge
Page 552

T able 9-3. Cables (continued)

Ref.

Desig.

W56 08360-20208 1 AMP MUL T/DUAL MOD{A28J3/A38J15

W56 08360-20258 1 AMP MUL T/AM MOD{A28J3/A38J8 (8360 B-Series Option 006),

W57 08360-20203 1 DUAL MOD/AMP FIL TER{A38J16/A29J2

W57 08360-20203 1 AM MOD/AMP FIL TER{A38J9/A29J2 (8360 B-Series Option 006)

W58 08360-20218 1 FIL TER/AMP SWITCH{A29J3/A33J2 (83623B )

W58 08360-20221 1 FIL TER/AMPLIFIER{A29J3/A33J2 (83624B)

W58 08360-20137 1 AMP FIL TER/DOUBLER{A29J3/A32J2 (83640B /L and 83650B/L)

W58 08360-20137 1 AMP FIL TER/DC BLOCKING CAP A CITOR{A29J3/A40J1

(83623L)

W58 08360-20250 1 AMP FIL TER/PLS MOD SW FIL T{A29J3/A42J5 (8360 B-Series

W59 08360-20078 1 AMP SWITCH/DIR COUPLER{A33J4/A30J1 (83623B )

W59 08360-20070 1 AMP SWITCH/DIR COUPLER{A33J4/A30J1

W59 08360-20107 1 AMPLIFIER/DIR COUPLER{A33J3/A30J1 (83624B Option 001)

W59 08360-20108 1 AMPLIFIER/DIR COUPLER{A33J3/A30J1 (83624B)

W59 08360-20194 1 DOUBLER/BRIDGE DET{A32J4/A30J4 (83640B/L and 83650B/L)

W59 08360-20192 1 DOUBLER/BRIDGE DET{A32J4/A30J4 (83640B/L and 83650B/L

W60 NOT USED

W61 NOT USED

W62 08360-60176 1 PULSE VIDEO OUT{A8J4/J13 (8360 B-Series Option 002)

W63 08360-60177 1 PULSE SYNC OUT{A8J6/J12 (8360 B-Series Option 002)

W64 08360-60180 1 AM/FM OUTPUT{A8J11/J11 (8360 B-Series Option 002)

W65 08360-60181 1 40 MHZ OUT{A8J21/A6J1 (8360 B-Series Option 002)

9-38 Replaceable P arts Agilent 8360

Show 6 more labels / short text on this page
P art · Qty Description · Num b er · Option 006) · (83623B Option 001) · Option 001)
Page 553 of the Agilent Technologies 83640B user manual — W73 08360-60250 1 RF INTERF A CE/BRIDGE DETECTOR{A20J13/A30J3
Page 553 · click to enlarge
Page 553

T able 9-3. Cables (continued)

Ref.

Desig.

W66 08360-60183 1 PULSE OUT{A8J5/A9J4 (8360 B-Series Option 002)

W67 08360-60184 1 AM OUT{A8J8/A10J1 (8360 B-Series Option 002)

W68 08360-60185 1 FM OUT{A8J9/A11J1 (8360 B-Series Option 002)

W69 08360-60224 1 RF INTERF A CE/DIODE SWITCH{A20J12/A41J1 (All mo dels except

83624B)

W70 08360-60248 1 F AST PULSE HI{A9J3/A42J1 (8360 B-Series Option 006)

W71 08360-20247 1 PLS MOD SW FIL T/DOUBLER{A42J6/A32J2

W71 08360-20252 1 PLS MOD SW FIL T/BRIDGE DET{A42J6/A30J4

W71 08360-20249 1 PLS MOD SW FIL T/COUPLER{A42J6/A30J1 (83623B/24B Options

W71 08360-20251 1 PLS MOD SW FIL T/BRIDGE DET{A42J6/A30J4

(83620B/22B/30B Options 001 and 006)

W71 08360-20248 1 PLS MOD SW FIL T/COUPLER{A42J6/A30J1 (83623B/24B Option

W72 08360-60191 1 RF INTERF A CE/PLS MOD SW FIL T{A20J11/A42J8 (8360 B-Series

W73 08360-60250 1 RF INTERF A CE/BRIDGE DETECTOR{A20J13/A30J3

Agilent 8360 Replaceable P arts 9-39

Show 8 more labels / short text on this page
P art · Qty Description · Num b er · (83640B/50B Option 006) · (83620B/22B/30B Option 006) · 001 and 006) · 006) · Option 006)
Page 554 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 554 · click to enlarge
Page 554

Front P anel Hardw are

Agilent Technologies 83640B — User Cal Menu diagram (page 554)
Fig. 217 · User Cal Menu

Figure 9-4. Front P anel Hardw are

9-40 Replaceable P arts Agilent 8360

Page 555 of the Agilent Technologies 83640B user manual — 2950-0043 1 NUT-HEX-DBL-CHAM 3/8-32-THD .094-IN-THK
Page 555 · click to enlarge
Page 555

T able 9-4. Front P anel Hardw are

Ref.

Desig.

1 0515-2043 7 SCREW-MA CH. M4 x 0.7 8MM-LG 90-DEG-FLH-HD

2 0515-0943 2 SCREW-MA CH. M4 x 0.7 12MM-LG

0535-0082 2 NUT-SPECIAL TY M3.5 x 1.27 THD; 8.44MM

3 0370-3033 1 R OT AR Y KNOB

4 0515-2043 SCREW-MA CH. M4 x 0.7 8MM-LG 90-DEG-FLH-HD

5 0515-0943 SCREW-MA CH. M4 x 0.7 12MM-LG

0535-0082 2 NUT-SPECIAL TY M3.5 x 1.27 THD; 8.44MM

6 4 J2 FL TG BNC PLS GRD

4 (Includes the following)

2190-0102 W ASHER-LK INTL T 15/32 IN .472-IN-ID

0590-1251 NUT-SPCL Y 15/32-32-THD .1-IN-THK .562-WD

7 J3 FL TG BNC A Y-FM (8360 B-Series)

2190-0102 W ASHER-LK INTL T 15/32 IN .472-IN-ID

0590-1251 NUT-SPCL Y 15/32-32-THD .1-IN-THK .562-WD

8 J4 FL TG BNC A Y-AM (8360 B-Series)

2190-0102 W ASHER-LK INTL T 15/32 IN .472-IN-ID

0590-1251 NUT-SPCL Y 15/32-32-THD .1-IN-THK .562-WD

9 J5 FL TG BNC A Y-EXT (8360 B-Series)

2190-0102 W ASHER-LK INTL T 15/32 IN .472-IN-ID

0590-1251 NUT-SPCL Y 15/32-32-THD .1-IN-THK .562-WD

10 08360-40007 1 DISPLA Y FIL TER

11 1251-5436 4 SCREW LOCK KIT-AMP HD CONN

J1 RF OUTPUT CONNECTOR - See T able 9-2

RPG1 08340-60326 1 R OT AR Y PULSE GENERA TOR (RPG)

2190-0016 1 W ASHER-LK INTL T 3/8 IN .377-IN-ID

2950-0043 1 NUT-HEX-DBL-CHAM 3/8-32-THD .094-IN-THK

Agilent 8360 Replaceable P arts 9-41

Show 7 more labels / short text on this page
P art · Qty Description · Num b er · (Includes the following) · (Includes the following) · (Includes the following) · (Includes the following)
Page 556 of the Agilent Technologies 83640B user manual — 1 08360-60146 1 KEYP AD REPLA CEMENT KIT
Page 556 · click to enlarge
Page 556

Front P anel, Casting and Keypad

Agilent Technologies 83640B — User Cal Menu diagram (page 556)
Fig. 218 · User Cal Menu

Figure 9-5. Front P anel, Casting and Keypad

T able 9-5. Front P anel, Casting and Keypad

Ref.

Desig.

1 08360-60146 1 KEYP AD REPLA CEMENT KIT

2 08360-20051 1 FRONT P ANEL CASTING - FEEDS (8360 B-Series only)

3 06360-20050 1 FRONT P ANEL CASTING - NO FEEDS (Option 004 for all 8360 B/L

4 06360-20287 1 FRONT P ANEL CASTING - (8360 L-Series only)

F or 2, 3, and 4, also order front panel nameplate from T able 9-16.

9-42 Replaceable P arts Agilent 8360

Show 4 more labels / short text on this page
P art · Qty Description · Num b er · mo dels)
Page 557 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 557 · click to enlarge
Page 557

Front P anel, A ttaching Hardw are

Agilent Technologies 83640B — User Cal Menu diagram (page 557)
Fig. 219 · User Cal Menu

Figure 9-6. Front P anel, A ttaching Hardw are (1 of 3)

T able 9-6. Front P anel, A ttaching Hardw are

Ref.

Desig.

1 0515-0372 27 SCREW-MA CHINE ASSEMBL Y M3 x 0.5 8MM-LG

2 0515-0664 2 SCREW-MA CHINE ASSEMBL Y M3 x 0.5 12MM-LG

3 08360-00034 2 MOUNTING BRA CKET - DISPLA Y

4 0515-0372 SCREW-MA CHINE ASSEMBL Y M3 X 0.5 8MM-LG

5 0515-0372 SCREW-MA CHINE ASSEMBL Y M3 X 0.5 8MM-LG

Agilent 8360 Replaceable P arts 9-43

P art · Qty Description · Num b er

Page 558 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 558 · click to enlarge
Page 558

Front P anel, A ttaching Hardw are

Agilent Technologies 83640B — User Cal Menu diagram (page 558)
Fig. 220 · User Cal Menu

Figure 9-6. Front P anel, A ttaching Hardw are (2 of 3)

9-44 Replaceable P arts Agilent 8360

Page 559 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 559 · click to enlarge
Page 559

Front P anel, A ttaching Hardw are

Agilent Technologies 83640B — User Cal Menu diagram (page 559)
Fig. 221 · User Cal Menu

Figure 9-6. Front P anel, A ttaching Hardw are (3 of 3)

Agilent 8360 Replaceable P arts 9-45

Page 560 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 560 · click to enlarge
Page 560

Rear P anel Hardw are

Agilent Technologies 83640B — User Cal Menu diagram (page 560)
Fig. 222 · User Cal Menu

Figure 9-7. Rear P anel Hardw are

9-46 Replaceable P arts Agilent 8360

Page 561 of the Agilent Technologies 83640B user manual — 10 0380-0643 2 ST ANDOFF HEX .255-IN-LG 6/32-THD
Page 561 · click to enlarge
Page 561

Rear P anel Hardw are

T able 9-7. Rear P anel (1 of 2)

Ref.

Desig.

1 08360-00001 1 REAR P ANEL

2 0515-1372 5 SCREW-MA CHINE ASSEMBL Y M3 x 0.5 8MM-LG

3 BNC CONNECTORS (Order A34 of T able 9-2 )

6960-0027 4 PLUG-HOLE TR-HD FOR .625-D-HOLE NYL

4 08360-60094 1 J8 - CA A Y-PLS INPUT (Option 004)

6960-0041 1 PLUG-HOLE TR-HD FOR .5-D-HOLE NYL (Standard)

5 J6 - CA A Y-10MHZ IN-See T able 9-3

2950-0035 4 NUT-HEX DBL-CHAM 15/32-32-THD

2190-0102 4 W ASHER LK INTL T 15/32 IN .472-IN-ID

6 J10 - CA A Y-V OL TS/GHZ-See T able 9-3

2950-0035 NUT-HEX DBL-CHAM 15/32-32-THD

2190-0102 W ASHER LK INTL T 15/32 IN .472-IN-ID

7 J7 - CA A Y-10MHZ OUT-See T able 9-3

2950-0035 NUT-HEX DBL-CHAM 15/32-32-THD

2190-0102 W ASHER LK INTL T 15/32 IN .472-IN-ID

8 J9 - CA A Y-ANAL INTFC-See T able 9-3

2950-0035 NUT-HEX DBL-CHAM 15/32-32-THD

2190-0102 W ASHER LK INTL T 15/32 IN .472-IN-ID

9 0515-0380 8 SCREW-MA CHINE ASSEMBL Y M4 x 0.7 10MM-LG

10 0380-0643 2 ST ANDOFF HEX .255-IN-LG 6/32-THD

11 1251-5436 4 SCREW LOCK KIT-AMP HD CONN

12 0515-1037 2 SCREW-MA CHINE ASSEMBL Y M3 x 0.5 45MM-LG

13 0515-0372 9 SCREW-MA CHINE ASSEMBL Y M3 x 0.5 8MM-LG

14 LINE MODULE - FIL TERED (See T able 9-2 )

15 08360-60032 1 A C V OL T A GE SELECTION SWITCH/HARNESS

Agilent 8360 Replaceable P arts 9-47

Show 7 more labels / short text on this page
P art · Qty Description · Num b er · (Includes the following) · (Includes the following) · (Includes the following) · (Includes the following)
Page 562 of the Agilent Technologies 83640B user manual — 2110-0564 1 FUSEHOLDER BOD Y
Page 562 · click to enlarge
Page 562

T able 9-7. Rear P anel (2 of 2)

Ref.

Desig.

16 F AN (see T able 9-2 )

17 08360-00025 1 PLA TE - REAR BLANK

6960-0027 4 PLUG-HOLE TR-HD FOR .625-D-HOLE NYL

18 1 J11-CA A Y-AM/FM OUTPUT (Option 002)

2950-0035 NUT-HEX DBL-CHAM 15/32-32-THD

2190-0102 W ASHER LK INTL T 15/32 IN .472-IN-ID

6960-0041 1 PLUG-HOLE TR-HD FOR .5-D-HOLE NYL (Standard)

19 1 J12-CA A Y-PULSE SYNC OUT (Option 002)

2950-0035 NUT-HEX DBL-CHAM 15/32-32-THD

2190-0102 W ASHER LK INTL T 15/32 IN .472-IN-ID

6960-0041 1 PLUG-HOLE TR-HD FOR .5-D-HOLE NYL (Standard)

20 1 J13-CA A Y-PULSE VIDEO OUT (Option 002)

2950-0035 NUT-HEX DBL-CHAM 15/32-32-THD

2190-0102 W ASHER LK INTL T 15/32 IN .472-IN-ID

6960-0041 1 PLUG-HOLE TR-HD FOR .5-D-HOLE NYL (Standard)

21 1250-1753 1 J14-CONNECTOR-A UX OUTPUT

22 08360-20245 1 REAR P ANEL INTERF A CE SHIELD

23 0515-2032 2 SCREW-MA CH M3 x 0.5 8MM-LG 90-DEG-FLH-HD

A T1 1810-0118 1 TERMINA TION SMA MALE 50

F1 2110-0227 1 FUSE (INCH) 5A 250V NTD FE UL

2110-0003 1 FUSE (INCH) 3A 230V NTD FE UL

2110-0564 1 FUSEHOLDER BOD Y

2110-0565 1 FUSEHOLDER CAP

2110-0569 1 FUSEHOLDER NUT

9-48 Replaceable P arts Agilent 8360

Show 6 more labels / short text on this page
P art · Qty Description · Num b er · (Includes the following) · (Includes the following) · (Includes the following)
Page 563 of the Agilent Technologies 83640B user manual — 10 08360-00061 2 BRA CKET F AN
Page 563 · click to enlarge
Page 563

Rear P anel F an Assembly

Agilent Technologies 83640B — User Cal Menu diagram (page 563)
Fig. 223 · User Cal Menu

Figure 9-8. Rear P anel F an Assembly

T able 9-8. Rear P anel F an Assembly

Ref.

Desig.

1 0515-0430 2 SMM 3.0 6 PM TX

2 08360-00059 1 F AN SHR OUD

3 08360-20254 1 FO AM

4 08360-00058 1 HEX CELL HONEYCOMB

5 08360-40011 1 GR OUND HEX SHR OUD

6 08360-00060 1 REAR P ANEL

7 0515-2011 8 SCREW-MA CHINE ASSY M3.5 x 0.6

8 0400-0356 4 GR OMMET

9 08360-00057 2 BRA CKET F AN/REAR P ANEL

10 08360-00061 2 BRA CKET F AN

11 08360-80026 1 F AN

Agilent 8360 Replaceable P arts 9-49

P art · Qty Description · Num b er

Page 564 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 564 · click to enlarge
Page 564

Bottom View, A ttaching Hardw are

Agilent Technologies 83640B — User Cal Menu diagram (page 564)
Fig. 224 · User Cal Menu

Figure 9-9. Bottom View, A ttaching Hardw are

T able 9-9. Bottom View, A ttaching Hardw are

Ref.

Desig.

1 6960-0016 2 PLUG-HOLE TR-HD FOR .125-D HOLE NYL

2 SCREW-MA CHINE (See T able 9-7 )

3 0624-0518 3 SCREW-TPG 6-19 .25-IN-LG P AN-HD-TORX T15

4 08360-40002 1 HIGH V OL T A GE CO VER

5 0515-0382 16 SCREW-MA CHINE ASSEMBL Y M4 x 0.7 12MM-LG

6 SCREW-MA CHINE (See T able 9-4 )

7 SCREW-MA CHINE (See T able 9-4 )

9-50 Replaceable P arts Agilent 8360

P art · Qty Description · Num b er

Page 565 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 565 · click to enlarge
Page 565

T op View, A ttaching Hardw are

Agilent Technologies 83640B — User Cal Menu diagram (page 565)
Fig. 225 · User Cal Menu

Figure 9-10. T op View, A ttaching Hardw are

Agilent 8360 Replaceable P arts 9-51

Page 566 of the Agilent Technologies 83640B user manual — 23 0515-0430 2 SMM 3.0 6 PN TX
Page 566 · click to enlarge
Page 566

T able 9-10. T op View, A ttaching Hardw are

Ref.

Desig.

1 08360-00015 1 PO WER SUPPL Y HOUSING

2 SCREW-MA CHINE (See T able 9-8 )

3 08360-20044 1 END SUPPOR T

4 0515-0501 2 SCREW-SKT-HD-CAP M6 x 1.0 12MM-LG

5 5021-7872 1 MOUNTING BRA CKET BRIDGE DETECTOR

(83620B/22B/30B/40B/50B)

5 08360-00008 1 MOUNTING BRA CKET COUPLER/DETECTOR (83623B /24B)

6 0515-0372 2 SCREW-MA CHINE ASSEMBL Y M3 x 0.5 8MM-LG (Quan tity is 4 in

instruments with Option 001)

7 0520-0173 2 SCREW-MA CH 256 .188 PNPD

8 0515-0430 2 SMM 3.0 6 PN TX

9 08360-00035 1 SUPPOR T CONNECTORS

10 08360-20237 1 EXTR USION-FIL TER MOUNT

0515-0383 1 SCREW-MA CHINE ASSEMBL Y M4 x 0.7 16MM-LG

11 08360-20042 1 CARD CA GE SUPPOR T

12 SCREW-MA CHINE (See T able 9-4 )

13 SCREW-MA CHINE (See T able 9-4 )

14 08360-20043 3 DIVIDERS

15 8160-0539 RFI R OUND STRIP CU-CLD-STL SN-PL (2.56 FT)

16 0515-0382 6 SCREW-MA CHINE ASSEMBL Y M4 x 0.7 12MM-LG

17 08360-40004 RFI SEAL (3 FT)

18 08360-20045 1 PO WER SUPPL Y CO VER

19 2200-0103 2 SCREW-MA CH. 4-40 .25-IN-LG P AN-HD-POZI

2190-0105 2 W ASHER-LK HLCL NO. 6 .141-IN-ID

1520-0205 3 SHOCK MOUNT .31-EFF-HGT

20 0515-1539 4 SCREW-MA CHINE M3 x0.5 16MM-LG

21 08360-00052 1 MOUNTING BRA CKET COUPLER/DIODE SWITCH

22 08360-00031 1 MOUNTING BRA CKET STEP A TTENUA TOR

23 0515-0430 2 SMM 3.0 6 PN TX

9-52 Replaceable P arts Agilent 8360

P art · Qty Description · Num b er

Page 567 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 567 · click to enlarge
Page 567

Left Side View, A ttaching Hardw are

Agilent Technologies 83640B — User Cal Menu diagram (page 567)
Fig. 226 · User Cal Menu

Figure 9-11. Left Side View, A ttaching Hardw are

T able 9-11. Left Side View, A ttaching Hardw are

Ref.

Desig.

1 0515-2086 4 SCREW-SPCL M4 x 0.7 7MM-LG 90-DEG-FLH-HD

2 0515-0382 15 SCREW-MA CHINE ASSEMBL Y M4 x 0.7 12MM-LG

3 0515-2086 SCREW-SPCL M4 x 0.7 7MM-LG 90-DEG-FLH-HD

Agilent 8360 Replaceable P arts 9-53

P art · Qty Description · Num b er

Page 568 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 568 · click to enlarge
Page 568

Right Side View, A ttaching Hardw are

Agilent Technologies 83640B — User Cal Menu diagram (page 568)
Fig. 227 · User Cal Menu

Figure 9-12. Right Side View, A ttaching Hardw are

T able 9-12. Right Side View, A ttaching Hardw are

Ref.

Desig.

1 0515-0382 33 SCREW-MA CHINE ASSEMBL Y M4 x 0.7 12MM-LG

2 0515-2086 8 SCREW-SPCL M4 x 0.7 7MM-LG 90-DEG-FLH-HD

3 0515-0433 2 SCREW-MA CHINE ASSEMBL Y M4 x 0.7 8MM-LG

4 0515-0382 SCREW-MA CHINE ASSEMBL Y M4 x 0.7 12MM-LG

5 0515-2086 SCREW-SPCL M4 x 0.7 7MM-LG 90-DEG-FLH-HD

9-54 Replaceable P arts Agilent 8360

P art · Qty Description · Num b er

Page 569 of the Agilent Technologies 83640B user manual — 6 08360-20141 1 RF DECK CASTING
Page 569 · click to enlarge
Page 569

RF Deck Right Side View, A ttaching Hardw are

Agilent Technologies 83640B — User Cal Menu diagram (page 569)
Fig. 228 · User Cal Menu

Figure 9-13. RF Deck Right Side View, A ttaching Hardw are

T able 9-13. RF Deck Right Side View, A ttaching Hardw are

Ref.

Desig.

1 0515-0372 17 SCREW-MA CHINE ASSEMBL Y M3 x 0.5 8MM-LG

2 3050-0169 4 W ASHER-SPR CR VD NO.6 .143-IN-ID

3 0515-2011 4 SCREW-MA CHINE ASSEMBL Y M3.5 x 0.060 12MM-LG

4 2360-0370 4 SCREW-MA CH 6-32 .375-IN-LG P AN-HD-POZI

5 0515-0372 SCREW-MA CHINE ASSEMBL Y M3 x 0.5 8MM-LG

6 08360-20141 1 RF DECK CASTING

7 0515-0372 SCREW-MA CHINE ASSEMBL Y M3 x 0.5 8MM-LG

8 0515-0458 2 SCREW-MA CHINE ASSEMBL Y M3 x 0.6 8MM-LG (Option 006)

9 0515-0430 2 SMM3.0 6 PN TX (Used for bridge detector regardless of its p osition on

Agilent 8360 Replaceable P arts 9-55

Show 4 more labels / short text on this page
P art · Qty Description · Num b er · the RF deck)
Page 570 of the Agilent Technologies 83640B user manual — 4 2190-0016 1 W ASHER-LK INTL 3/8 IN
Page 570 · click to enlarge
Page 570

RF Connector and A ttaching Hardw are

Agilent Technologies 83640B — User Cal Menu diagram (page 570)
Fig. 229 · User Cal Menu

Figure 9-14. RF Connector and A ttaching Hardw are

T able 9-14. RF Connector and A ttaching Hardw are

Ref.

Desig.

1 08360-20095 1 FRONT MOUNTING BRA CKET 3.5 MM CONNECTOR

08360-20096 1 FRONT MOUNTING BRA CKET 2.4 MM CONNECTOR

2 08360-20088 1 SP A CER FOR 3.5 MM CONNECTOR

3 2950-0001 1 NUT-HEX-DBL-CHAM 3/8-32-THD (3.5 mm connector)

08360-20133 1 NUT-HEX-DBL-CHAM M10 X1 (2.4 mm connector)

4 2190-0016 1 W ASHER-LK INTL 3/8 IN

5 08360-00004 1 PL-RR CONN 3.5 (Option 004)

08360-00024 1 PL-RR CONN 2.4 (Option 004)

J1 RF CONNECTOR (See T able 9-2 )

9-56 Replaceable P arts Agilent 8360

P art · Qty Description · Num b er

Page 571 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 571 · click to enlarge
Page 571
Agilent Technologies 83640B — User Cal Menu diagram (page 571)
Fig. 230 · User Cal Menu

Figure 9-15. Chassis P arts

Agilent 8360 Replaceable P arts 9-57

Chassis P arts

Page 572 of the Agilent Technologies 83640B user manual — 1 5062-3736 1 CO VER-TOP
Page 572 · click to enlarge
Page 572

T able 9-15. Chassis P arts

Ref.

Desig.

1 5062-3736 1 CO VER-TOP

2 0515-1245 1 SCREW - CO VER

3 5021-5838 2 SIDE RAILS

4 5041-8802 1 TRIM STRIP

5 5021-8405 1 FRONT FRAME

6 5062-3800 2 HANDLE ASSEMBL Y

7 0515-0943 4 HANDLE SCREWS

8 5021-8497 2 TRIM FRONT HANDLE

9 5041-8801 4 FOOT -FM .5M

10 5062-3748 1 CO VER-BOTTOM

11 1460-1345 2 TIL T ST AND

12 5062-3843 2 CO VER-SIDE

13 5041-8819 2 CAP-FR ONT

14 0515-0707 4 STRAP SCREW

15 5041-8820 2 CAP-REAR

16 5062-3705 2 STRAP HANDLE

17 08360-20040 1 SIDE FRAME

18 5021-5806 1 REAR FRAME

19 54110-40502 4 FOOT REAR

20 0515-1444 4 SCREW - REAR FOOT

21 08360-20253 1 INSULA TED BOTTOM CO VER

9-58 Replaceable P arts Agilent 8360

P art · Qty Description · Num b er

Page 573 of the Agilent Technologies 83640B user manual — 5.5 MM NUT DRIVER
Page 573 · click to enlarge
Page 573

T able 9-16. Miscellaneous Replaceable Accessories

Ref.

Desig.

LABELS

7120-3737 1 LBL WRN HI V OL T

83620-80004 1 NAMEPLA TE 83620B

83621-80002 1 NAMEPLA TE 83621B

83622-80002 1 NAMEPLA TE 83622B

83623-80003 1 NAMEPLA TE 83623B

83624-80002 1 NAMEPLA TE 83624B

83630-80002 1 NAMEPLA TE 83630B

83631-80002 1 NAMEPLA TE 83631B

83640-80003 1 NAMEPLA TE 83640B

83650-80002 1 NAMEPLA TE 83650B

83651-80002 1 NAMEPLA TE 83651B

83623-80004 1 NAMEPLA TE 83623L

83630-80003 1 NAMEPLA TE 83630L

83640-80004 1 NAMEPLA TE 83640L

83650-80003 1 NAMEPLA TE 83650L

SER VICE TOOLS

T-10 TORX SCREWDRIVER

T-15 TORX SCREWDRIVER

#1 x 3 1/8 IN. POZIDRIVE SCREWDRIVER

#2 x 4 IN. POZIDRIVE SCREWDRIVER

5.5 MM NUT DRIVER

7 MM NUT DRIVER

.050 HEX DRIVER

9/16 IN. NUT DRIVER

5/16 IN OPEN END WRENCH

14MM OPEN END WRENCH

1251-2170 CO AX EXTRA CTOR TOOL-SUBMIN D CONNECTOR

08360-60065 SUBSTITUTE FRONT P ANEL

Agilent 8360 Replaceable P arts 9-59

P art · Qty Description · Num b er

Page 574 of the Agilent Technologies 83640B user manual — SOFTW ARE
Page 574 · click to enlarge
Page 574

T able 9-16. Miscellaneous Replaceable Accessories (continued)

Ref.

Desig.

08360-60060 TOOL KIT

08360-60086 EXTENDER BO ARD-DIGIT AL BO ARDS

08360-60085 EXTENDER BO ARD-ANALOG BO ARDS

08360-60081 EXTENDER BO ARD-POST REGULA TOR

08360-60077 EXTENDER BO ARD-SWITCHING REGULA TOR

08360-60030 EXTENDER BO ARD-RECTIFIER/FIL TER

08360-60084 EXTENDER BO ARD-RF DECK

08360-20286 RF TEST CABLE-FIL TER TO DIRECTIONAL COUPLER

08360-60108 EXTENDER CABLE-PO WER SUPPL Y

8710-1787 5 MM HEX BALL DRIVER

ACCESSORIES

TOUCH-UP P AINT

6010-1146 DO VE GRA Y - FOR FRAME AR OUND FRONT P ANEL & P AINTED

POR TIONS OF FRONT HANDLES

6010-1147 FRENCH GRA Y - FOR SIDE, TOP , & BOTTOM CO VERS

6010-1148 P AR CHMENT GRA Y - FOR RA CK MOUNT FLANGES, RA CK

SUPPOR T SHEL VES, & FRONT P ANEL

5041-9148 PC BO ARD EXTRA CTORS

5062-7202 CABLE SMI to 8349B

ADAPTERS

1250-2187 AD APTER 2.4 MM (F) TO K (F)

1250-2188 AD APTER 2.4 MM (F) TO 2.4 MM (F)

1250-1745 AD APTER TYPE-N TO 3.5 MM (F)

5061-5311 AD APTER 3.5 MM (F) TO 3.5 MM (F)

SOFTW ARE

08360-10001 1 DISK - SER VICE SUPPOR T SOFTW ARE

9-60 Replaceable P arts Agilent 8360

Show 4 more labels / short text on this page
P art · Qty Description · Num b er · (Includes the following)
Page 575 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 575 · click to enlarge
Page 575

T able 9-16. Miscellaneous Replaceable Accessories (continued)

Ref.

Desig.

Do cumen tation F or 8360 B-Series Sw ept Signal Generator

08360-90128 8360 B-Series Man ual Set {

(This includes the following four man uals:)

08360-90126 Quick Reference Guide

08360-90124 T roublesho oting Guide

08360-90179 Obsolete Blo ck Diagrams (not include d in manual set )

Do cumen tation F or 8360 L-Series Sw ept CW Generator

08360-90135 8360 L-Series Man ual Set {

(This includes the following four man uals:)

08360-90126 Quick Reference Guide

08360-90124 T roublesho oting Guide

08360-90179 Obsolete Blo ck Diagrams (not include d in manual set )

Do cumen tation F or 83621B/31B/51B

08360-90138 83621B/31B/51B Man ual Set

(This includes the following man uals:)

08360-90136 T roublesho oting Guide

Agilent T ec hnologies 8360 Comp onen t-Lev el Repair Man ual

08360-90141 Agilent T ec hnologies 8360 Comp onent-Level Repair Man ual

Agilent 8360 Replaceable P arts 9-61

Show 10 more labels / short text on this page
P art · Qty Description · Num b er · F or 83620B/22B/23B/24B/30B/40B/50B · 08360-90127 User's Guide · 08360-90125 Service Guide · F or 83623L/30L/40L/50L · 08360-90134 User's Guide · 08360-90125 Service Guide · 08360-90137 Service Guide
Page 576 of the Agilent Technologies 83640B user manual — 08360-60235 DEL OPTION 004 RETR OFIT KIT -
Page 576 · click to enlarge
Page 576

T able 9-16. Miscellaneous Replaceable Accessories (continued)

Ref.

Desig.

UPGRADE/OPTION RETR OFIT KITS

08360-60218 ADD OPTION 001 RETR OFIT KIT - 83620B/22B/24B, 83623B/L, and

83630B/L

08360-60220 ADD OPTION 001 RETR OFIT KIT - 83640B/L and 83650B/L

08360-60129 DEL OPTION 001 RETR OFIT KIT - 83620B/22B/24B, 83623B/L, and

83630B/L

08360-60221 DEL OPTION 001 RETR OFIT KIT - 83640B/L and 83650B/L

83602A ADD OPTION 002 RETR OFIT KIT (8360 B- Series)

08360-60236 ADD OPTION 004 RETR OFIT KIT - 83620B/22B/24B, 83623B/L, and

83630B/L

08360-60237 ADD OPTION 004 RETR OFIT KIT - 83640B/L and 83650B/L

08360-60235 DEL OPTION 004 RETR OFIT KIT -

83620B/22B/23B/24B/30B

08360-60238 DEL OPTION 004 RETR OFIT KIT -

83640B/50B

08360-60141 ADD OPTION 008 RETR OFIT KIT

08360-60142 ADD OPTION 700 RETR OFIT KIT

08360-60143 ADD OPTION 806 RETR OFIT KIT

08360-60144 ADD OPTION 908 RETR OFIT KIT

08360-60145 ADD OPTION 913 RETR OFIT KIT

08360-60201 FIRMW ARE UPGRADE KIT

9-62 Replaceable P arts Agilent 8360

P art · Qty Description · Num b er

Page 577 of the Agilent Technologies 83640B user manual — W ARNING
Page 577 · click to enlarge
Page 577

Prev entiv e Maintenance

This c hapter con tains the follo wing information on care and maintenance of the syn thesizer:

Ho w to Clean the F an Filter

Ho w to Clean the Displa y Filter

Ho w to Replace the Line F use

Ho w to Replace the CPU Battery

W ARNING

This is a Safety Class I product (pro vided with a protectiv e earthing ground

incorporated in the po w er cord). The mains plug shall only be inserted in a

sock et outlet pro vided with a protectiv e earth contact. Any interruption of the

protectiv e conductor, inside or outside the instrument, is lik ely to mak e the

instrument dangerous. Intentional interruption is prohibited.

Agilent 8360 Prev entiv e Maintenance 10-1

Introduction · Connector Care

Page 578 of the Agilent Technologies 83640B user manual — W ARNING
Page 578 · click to enlarge
Page 578

How to Clean the F an Filter

The co oling fan, lo cated on the rear panel, has a thin foam lter. Ho w often the lter m ust b e

cleaned dep ends on the en vironment in whic h the syn thesizer op erates. As the lter collects

dust, the fan sp eed increases to maintain air ow (as the fan sp eed increases, so do es the fan

noise). If the lter con tinues to collect dust after the fan reac hes maximum sp eed, air ow is

reduced and the syn thesizer's in ternal temp erature increases. If the in ternal temp erature

C, the syn thesizer automatically turns o and the amb er standb y LED turns on.

W ARNING

The detachable po w er cord is the instrument disconnecting device. It

disconnects the mains circuits from the mains supply before other parts of the

instrument. The front panel switch is only a standb y switch and is not a LINE

switch.

Clean the fan lter as follows:

1. T urn o the syn thesizer.
2. Remo v e the A C line cord.
3. Remo v e the tw o screws holding the fan shroud to the rear panel. (See Figure 10-1.)
4. Separate the fan shroud, the foam, and the honeycom b from the ground hex shroud.
5. Rinse the foam and honeycom b in w arm w ater, then dry .
6. Rev erse the remov al pro cedure to reassemble the syn thesizer.

10-2 Prev entiv e Maintenance Agilent 8360

reac hes 90 ·

Page 579 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 579 · click to enlarge
Page 579
Agilent Technologies 83640B — User Cal Menu diagram (page 579)
Fig. 231 · User Cal Menu

Figure 10-1. Remo ving the F an Filter

Agilent 8360 Prev entiv e Maintenance 10-3

Page 580 of the Agilent Technologies 83640B user manual — CA UTION
Page 580 · click to enlarge
Page 580

How to Clean the Displa y Filter

The syn thesizer's display is protected b y a plastic display lter. T o clean the display lter, use

mild soap or detergen t and w ater, or a commercial window cleaner (ammonia do es not h urt

the plastic surface). Use a soft, lint-free cloth. Do not use abrasiv e cleaners, tissues, or pap er

tow els whic h can scratch the plastic.

Under normal op eration, y ou will need to clean only the front side of the display lter.

Occasionally , dep ending on the en vironment in whic h the syn thesizer op erates, y ou may wish

to clean the bac k side of the lter as follo ws:

1. T urn the syn thesizer to standb y (yellow LED on).
2. Remo v e the A C line cord.

W ARNING

The detachable po w er cord is the instrument disconnecting device. It

disconnects the mains circuits from the mains supply before other parts of the

instrument. The front panel switch is only a standb y switch and is not a LINE

switch.

CA UTION

When y ou remov e the front panel, if y ou do not disconnect the A C line cord

y ou can easily blow the fuse on the preregulator assembly .

3. Refer to Chapter 7 and remov e the front panel, front panel pro cessor, alphan umeric display ,

and the source mo dule in terface.

4. Clean the bac k side of the display lter and reassemble the instrumen t.

10-4 Prev entiv e Maintenance Agilent 8360

Page 581 of the Agilent Technologies 83640B user manual — W ARNING
Page 581 · click to enlarge
Page 581

How to Replace the Line Fuse

The line fuse is lo cated on the instrumen t rear panel. Replace this fuse as follows:

1. T urn the syn thesizer to standb y (yellow LED on).
2. Remo v e the A C line cord.

W ARNING

The detachable po w er cord is the instrument disconnecting device. It

disconnects the mains circuits from the mains supply before other parts of the

instrument. The front panel switch is only a standb y switch and is not a LINE

switch.

3. Refer to Figure 10-2. Using a small at{blade screwdriver, rotate the fuse cap

coun terclo c kwise, and remov e the fuse holder.

4. Replace the original fuse.

W ARNING

For continued protection against fire hazard replace line fuse only with same

type and rating (refer to Figure 10-2). The use of other fuses or material is

prohibited.

5. Replace the fuse holder in the rear panel. Using the screwdriver, rotate the fuse cap

clo c kwise to secure the fuse holder in place.

6. Reconnect the syn thesizer to line p ow er.
Agilent Technologies 83640B — W ARNING diagram (page 581)
Fig. 232

Figure 10-2. Replacing the Line Fuse

Agilent 8360 Prev entiv e Maintenance 10-5

Page 582 of the Agilent Technologies 83640B user manual — CA UTION
Page 582 · click to enlarge
Page 582

How to Replace the CPU Battery (A15BT1)

W ARNING

same or equiv alent type recommended.

Battery A15BT1 contains lithium iodide. Do not incinerate or puncture this

battery . Dispose of the discharged battery in a safe manner.

Do not thro w batteries a w ay but collect as small chemical w aste.

CA UTION

The A15 CPU Assem bly may b e damaged if it is placed on a conductiv e

surface. Use a static-safe w ork station. Ensure that the CPU Assem bly is not

placed on an y conductiv e material.

1. Remo v e the A15 CPU Assem bly . Refer to the \6. Ma jor Assem blies Remo v al" in

Chapter 7 pro cedure in Chapter 7.

Agilent Technologies 83640B — CA UTION diagram (page 582)
Fig. 233

Figure 10-3. A15BT1 Battery Location

CA UTION

Placing the b ottom side of the A15 assembly on a conductiv e w ork surface

without an insulator can result in loss of correction data stored in RAM.

Protect the A15 assembly from con tact with an y conductiv e surface.

2. Lo cate the battery leads on the A15 assembly , unsolder them, and remov e the old battery .
3. Replace and solder the leads on the new battery , ensuring prop er p olarit y as silkscreened

on the A15 assembly .

4. Record the battery-replacemen t date on the battery lab el.
5. Replace the A15 CPU Assem bly . Refer to \6. Ma jor Assem blies Remo v al" in Chapter 7.

10-6 Prev entiv e Maintenance Agilent 8360

Remo v al/Replacement

Page 583 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 583 · click to enlarge
Page 583

Clean and gage the syn thesizer's output connector p erio dically . F or details, refer to

Applic ation Note 326 , whic h do cumen ts the principles of microw av e connector care.

Agilent 8360 Prev entiv e Maintenance 10-7

Connector Care

Page 584 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 584 · click to enlarge
Page 584
Page 584 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 585 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 585 · click to enlarge
Page 585

This c hapter provides option retro t information. Eac h option and the complexity of the

retro t pro cedure are brie y describ ed. Some options cannot b e retro tted and so are not

men tioned in this c hapter. Where applicable, kit part n um b ers are provided for b oth adding

and deleting the option. The sp eci c details of the retro t, including an y adjustmen ts and

p erformance tests required after an assembly is replaced, are discussed in the installation

n um b er.

Option Numb er Option Name

001 Add A tten uator

002 Add Mo dulation/Generator

004 Rear P anel RF Output

008 1 Hz Capabilit y

700 MA TE Compatibilit y

806 Rack Moun t Side Kit

908 Rack Flange Kit Without Handles

910 Extra Man ual Set

913 Rack Flange Kit With Handles

Agilent 8360 Option Retrofits 11-1

Option Retrofits · Introduction · Option Retrofits

Page 586 of the Agilent Technologies 83640B user manual — 83622B
Page 586 · click to enlarge
Page 586

Add A ttenuator (Add Option 001)

Option 001 adds an attenuator to the RF dec k to provide the capabilit y of setting a

wider range of output p ow er. T o add the attenuator, the RF dec k m ust b e recon gured.

Adjustmen ts and p erformance tests that are required after installation of the attenuator

are listed in Chapter 8 and in the installation note. Because the step attenuator atness

test requires v ery sp eci c test equipmen t that is not part of the standard microw a v e test

system, and b ecause sev eral exp ensive micro circuits m ust b e handled in the pro cedure, it is

recommended that the instrumen t b e returned to an Agilent service cen ter for retro tting.

Retro t Kit P art Numb er

Mo del Numb er

83620B 08360-60218

83622B

83630B/L

83623B/L 08360-60252

83624B

83640B/L 08360-60220

83650B/L

Delete A ttenuator (Delete Option 001)

Retro t Kit P art Numb er

Mo del Numb er

83622B

83623B/L

83624B

83630B/L

83640B/L

83650B/L

11-2 Option Retrofits Agilent 8360

83620B None

Page 587 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 587 · click to enlarge
Page 587

Add Modulation/Generator (Add Option 002)

Option 002 adds a digitally-synthesized in ternal mo dulation source and AM depth and FM

deviation meters. Because this is a complex pro cedure, it is recommended that the instrumen t

b e returned to an Agilent service cen ter for retro tting. Purc hase of the retro t kit includes

installation at an Agilent service cen ter.

Retro t Kit Mo del Numb er

Mo del Numb er

All B-series mo dels Agilent 83602A

Agilent 8360 Option Retrofits 11-3

Page 588 of the Agilent Technologies 83640B user manual — 83620B 08360-60235
Page 588 · click to enlarge
Page 588

Rear P anel RF Output (Add Option 004)

Option 004 mov es the RF output, external ALC input, pulse input/output,

AM input, and FM input connectors from the front panel to the rear panel. Due to the

unique design of the RF dec k, there is no loss of output p ow er with the rear output option.

Adjustmen ts and p erformance tests that are required after installation are listed in Chapter 8

and in the installation note included in the retro t kit.

Retro t Kit P art Numb er

Mo del Numb er

83620B 08360-60236

83622B

83623B

83624B

83630B

83640B 08360-60237

83650B

Front P anel RF Output (Delete Option 004)

Retro t Kit P art Numb er

Mo del Numb er

83620B 08360-60235

83622B

83623B

83624B

83630B

83640B 08360-60238

83650B

11-4 Option Retrofits Agilent 8360

Page 589 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 589 · click to enlarge
Page 589

1 Hz Capability (Add Option 008)

Option 008 provides 1 Hz frequency resolution (1 kHz resolution is standard).

HP BASIC 5.1 and an HP 9000 series 200 or 300 computer are required to run the softw are to

retro t the instrumen t.

Retro t Kit P art Numb er

Mo del Numb er

All mo dels 08360-60141

MA TE Compatibility (Add Option 700)

Option 700 adds CI IL program commands to the instrumen t for MA TE system compatibilit y .

HP BASIC 5.1 and an HP 9000 series 200 or 300 computer are required to run the softw are to

retro t the instrumen t.

Retro t Kit P art Numb er

Mo del Numb er

All mo dels 08360-60142

Agilent 8360 Option Retrofits 11-5

Page 590 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 590 · click to enlarge
Page 590

Rack Mount Slide Kit (Add Option 806)

Option 806 con tains the necessary hardw are to moun t sliding rack moun ts on the syn thesizer.

This allo ws easier access to the syn thesizer when it is moun ted in an equipmen t rack. Y ou

m ust remov e the instrumen t side panels to install the kit. Instructions for installation are in

the installation note included in the retro t kit.

Retro t Kit P art Numb er

Mo del Numb er

All mo dels 08360-60143

Rack Flange Kit Without Handles (Add Option 908)

Option 908 con tains a pair of anges and the necessary hardw are to moun t the syn thesizer

( without hand les ) in an equipmen t rack with 482.6 mm (19 in.) horizontal spacing.

Instructions for installation are in the installation note included in the retro t kit.

Retro t Kit P art Numb er

Mo del Numb er

All mo dels 08360-60144

11-6 Option Retrofits Agilent 8360

Page 591 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 591 · click to enlarge
Page 591

Extra Manual Set (Add Option 910)

All instrumen ts are supplied with one complete man ual set whic h do cumen ts op eration and

service. Eac h Option 910 (av ailable only with the initial order) provides one additional

complete man ual set. T o order additional man uals after initial shipmen t, order the individual

man ual part n um b ers, or the man ual set part n um b ers listed in Chapter 9.

Rack Flange Kit With Handles (Add Option 913)

Option 913 con tains a pair of anges and the necessary hardw are to moun t the syn thesizer

( with hand les attache d ) in an equipmen t rack with 482.6 mm

(19 in.) horizontal spacing. Instructions for installation are in the installation note included in

the retro t kit.

Retro t Kit P art Numb er

Mo del Numb er

All mo dels 08360-60145

Agilent 8360 Option Retrofits 11-7

Page 592 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 592 · click to enlarge
Page 592
Page 592 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 593 of the Agilent Technologies 83640B user manual — PCO 3844A:11159
Page 593 · click to enlarge
Page 593

How to Use Instrument History

This man ual do cumen ts the curren t pro duction v ersions of the \standalone" Agilent

8360 B-Series Sw ept Signal Generator/ 8360 L-Series Sw ept CW Generator.

This man ual will b e mo di ed to apply to future v ersions of these instrumen t mo dels.

Information provided in this c hapter will then allo w y ou to adapt this man ual to the earlier

v ersions. If y our instrumen t is one of the earlier v ersions, y ou may ha v e to mo dify y our

man ual using the information in this c hapter. Chec k the serial n um b er pre x attac hed to y our

syn thesizer's rear panel and then lo cate it in the following tables. The tables tell y ou whic h

c hanges to make. Incorp orate the c hanges in rev erse alphab etical order.

Serial Numb er Pre x Required Changes

3844A No Change Needed

3722A and b elow A

Serial Numb er Pre x Required Changes

3844A No Change Needed

3722A and b elow A

Agilent T ec hnologies In ternal Use Only

PCO 3844A:11159

Show 5 more labels / short text on this page
Instrument History · Agilent 83620B/22B/23B/24B/30B/40B/50B · Agilent 83623L/30L/40L/50L · Instrument History 12-1 · Service
Page 594 of the Agilent Technologies 83640B user manual — W11 08360-60074 1 125 KHZ REF{A7J3/A4J1
Page 594 · click to enlarge
Page 594

The A4 Fractional-N Assembly has Changed

All Agilent 8360 B-Series and 8360 L-Series instrumen ts with serial pre x n um b ers 3722A and

b elow ha v e a F ractional-N assembly that is di eren t than the one do cumen ted in this man ual.

As a result, y ou m ust apply the c hanges describ ed in this section in order for this man ual to

conform to y our instrumen t.

The part n um b ers listed in T able 12-1 are correct for y our instrumen t. Use these part

n um b ers instead of the part n um b ers ha ving the same reference designators in the

\Replaceable P arts" c hapter. Refer to Figure 12-3 for assembly and cable lo cations.

T able 12-1.

Ref.

Desig.

A4 08360-60010 1 FRA CTIONAL{N ASSEMBL Y

A4 08360-69010 FRA CTIONAL{N ASSEMBL Y (Rebuilt-Exchange)

W11 08360-60074 1 125 KHZ REF{A7J3/A4J1

W12 08360-60075 1 30{60 MHZ{A4J2/A5J2

W19 08360-60096 1 10 MHZ REAR P ANEL {A7J5/J7

12-2 Instrument History Agilent 8360

Show 7 more labels / short text on this page
Change A · Change A · Replaceable P arts · P art · Qty Description · Num b er · Service
Page 595 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 595 · click to enlarge
Page 595
Agilent Technologies 83640B — User Cal Menu diagram (page 595)
Fig. 234 · User Cal Menu

Figure 12-3. Major Assembly and Cable Locations

Change A · Instrument History 12-3 · Service

Page 596 of the Agilent Technologies 83640B user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 596 · click to enlarge
Page 596

The following adjustmen ts apply to y our instrumen t. Use these adjustmen ts in addition to the

ones describ ed in the \Adjustmen ts" c hapter of this man ual.

Fractional-N V CO Adjustment

No test equipmen t is required for this pro cedure.

This adjustmen t sets the V CO tuning v oltage to 6.0 V for a V CO output of 60 MHz.

1. Lea ving all cables connected, place the A4 fractional-N assembly on an analog extender

b oard.

2. T urn the syn thesizer on and press

PRESET

. Let the instrumen t w arm up for at least one

hour.

3. Select the A4 V CO tune adjustmen t. On the syn thesizer, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SERVICE

4. Using a non-metallic to ol, adjust A4L1 (see Figure 12-4 ) to cen ter the needle on the

display .

Agilent Technologies 83640B — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 596)
Fig. 235

Figure 12-4. A4L1 Location

5. T urn the syn thesizer to standb y , and reinstall the A4 assembly .

Related P erformance T ests

12-4 Instrument History Agilent 8360

Show 10 more labels / short text on this page
Change A · Adjustments · Description and Procedure · · · Adjust Menu · AssyAdj Menu · A4 VCO Tune · Self-tests · Service
Page 597 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 597 · click to enlarge
Page 597

In Case of Difficulty

1. V erify that an analog extender b oard is used.
2. Make sure the 125 kHz reference cable, W11, is connected to A4J1.
3. Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal Gener ator/ 8360 L-Series

Swept CW Gener ator T r oublesho oting Guide .

Fractional-N Reference and API Spurs Adjustment

Do not p erform this adjustmen t on a new or rebuilt replacemen t assembly .

The A4 assembly is factory-adjusted, and is not instrumen t dep enden t.

This adjustmen t minimizes three kno wn spurs (125 kHz reference, API1, and API2). A t eac h

spur frequency , the sp ectrum analyzer is phase lo c k ed to the syn thesizer, connected directly to

the A4 fractional-N assembly output, and op erated in zero span. A4R70, R31, and R30 are

adjusted for minimum spur lev el. T able 12-2 summarizes this adjustmen t.

T able 12-2. Summary of Fractional-N Reference & API Spurs Adjustment

P oten tiometer Minimum

125 kHz Reference 44 .125 44 .0 A4R70 75

API 1 44 .005 44 .0 A4R31 70

API 2 44 .0005 44 .0055 A4R30 80

1. With the A4 assembly on an analog extender b oard, connect the equipmen t as sho wn in

Figure 12-5 . Preset the instrumen ts, and let them w arm up for at least one hour.

Show 11 more labels / short text on this page
Change A · Description and Procedure · Spur F ractional-N · Analyzer · CW F requency · F requency · dBc · (MHz) · (MHz) · Instrument History 12-5 · Service
Page 598 of the Agilent Technologies 83640B user manual — CW
Page 598 · click to enlarge
Page 598
Agilent Technologies 83640B — User Cal Menu diagram (page 598)
Fig. 236 · User Cal Menu

Figure 12-5. Fractional-N Reference and API Spur Adjustment Setup

Sp ectrum Analyzer HP/Agilent 8566B

2. On the sp ectrum analyzer, set:
Frequency Span0 Hz
Resolution Bandwidth10 Hz
Sweep Time10 s
Reference Level 10 dBm
Scale Log10 dB/Division
Video AveragingOff
Clear Write ASelected

125 kHz Reference Spur Adjustment

3. On the syn thesizer, press:

CW

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SERVICE

CW

44.125

4. On the sp ectrum analyzer, set:
Center Frequency44.125 MHz
Frequency Span500 Hz

Adjust the frequency o set to cen ter the signal on the display .

Frequency Span0 Hz

P eak the signal on the display .

12-6 Instrument History Agilent 8360

Show 14 more labels / short text on this page
Change A · Equipment · · · · · Tools Menu · CntlPLL Menu · Frac N Menu · · · MHz · · Service
Page 599 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 599 · click to enlarge
Page 599
Center Frequency44.0 MHz
Reference Level 40 dBm
5. On the syn thesizer, adjust A4R70 (see Figure 12-6) for minimum signal on the sp ectrum

analyzer.

Agilent Technologies 83640B — User Cal Menu diagram (page 599)
Fig. 237 · User Cal Menu

Figure 12-6. Fractional-N Reference and API Spurs Adjustment Locations

The di erence in lev el b et w een the signal noted in step 4 and the signal in this step should b e

at least 75 dBc.

Change A · Instrument History 12-7 · Service

Page 600 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 600 · click to enlarge
Page 600

API 1 Spur Adjustment

6. On the syn thesizer, c hange the fractional-N to CW 44.005 MHz. Press:

44.005

7. On the sp ectrum analyzer, set:
Center Frequency44.005 MHz
Reference Level+10 dBm
Center Frequency44.0 MHz
Reference Level 40 dBm
8. On the syn thesizer, adjust A4R31 (see Figure 12-6) for minimum signal on the sp ectrum

analyzer.

The di erence in lev el b et w een the signal noted in step 7 and the signal in this step

should b e at least 70 dBc.

API 2 Spur Adjustment

9. On the syn thesizer, c hange the fractional-N to CW 44.0005 MHz. Press

44.0005

10. On the sp ectrum analyzer, set:
Center Frequency44.0005 MHz
Reference Level+10 dBm
Center Frequency44.0055 MHz
Reference Level 40 dBm
11. On the syn thesizer, adjust A4R30 (see Figure 12-6) for minimum signal on the sp ectrum

analyzer.

The di erence in lev el b et w een the signal noted in step 10 and the signal in this step

should b e at least 80 dBc.

12. T urn the syn thesizer to standb y and reinstall the A4 assembly .

Related P erformance T ests

In Case of Difficulty

1. Ensure that the sp ectrum analyzer is lo c k ed to the external reference.
2. Y ou may need to use video av eraging on the sp ectrum analyzer if the spur lev el is low

enough to b e hidden b y phase noise.

3. If y ou cannot adjust spurs b elow the recommended lev els, the problem is probably with the

A4 fractional-N assembly . Refer to the A gilent T e chnolo gies 8360 B-Series Swept Signal

Gener ator/ 8360 L-Series Swept CW Gener ator T r oublesho oting Guide .

12-8 Instrument History Agilent 8360

Show 8 more labels / short text on this page
Change A · · · MHz · MHz · · Spurious Signals (Non-Harmonic) · Service
Page 601 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 601 · click to enlarge
Page 601

Use the information in T able 12-3 instead of the equiv alen t F ractional{N information provided

in the \Calibration Constants" c hapter.

T able 12-3.

Assemblies and Corresponding Cal Constant ID Numbers and Revisions

Assembly ID Numb er Revision Part Numb er Description

0 08360-60010 F ractional{N

Use the information in T able 12-4 instead of the equiv alen t F ractional{N information provided

in the \Post Repair" c hapter.

T able 12-4.

Adjustments and P erformance T ests Required

After Repair or Replacement of an Assembly

Assembly Adjustmen t Performance T est

F ractional{N V CO

F ull Self{T est

F requency Switching Time

The adjustmen ts listed are required only after r ep air of the assembly . The r eplac ement assembly

is adjusted at the factory and is not instrument-dep enden t.

Show 11 more labels / short text on this page
Change A · Calibration Constants · A4 None #496 · P ost Repair · A4 F ractional{N · F ractional{N Reference · and API Spurs · Spurious Signals · (Non-Harmonics) · Instrument History 12-9 · Service
Page 602 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 602 · click to enlarge
Page 602
Page 602 of the Agilent Technologies 83640B user manual — User Cal Menu
Page 603 of the Agilent Technologies 83640B user manual — BASIC
Page 603 · click to enlarge
Page 603

AM/FM DA C o set and gain adjustmen t, 3-73

10 MHz reference standard remo v al, 7-29

AM input imp edance adjustmen t, 3-86

AM o set adjustmen t, 3-78

adjustmen t interv al, 3-8

ampli er/ lter adjustmen t, 3-32

10 MHz standard adjustmen t, 3-4

ampli er lter remo v al, 7-44

125 kHz reference spur, 12-5

ampli er multiplier remo v al, 7-44

ampli er remo v al, 7-44

A

ampli er switc h remo v al, 7-44

A9 pulse b oard jump er adjustmen t, 3-97

ADC adjustmen t, 5-13

10 MHz standard, 3-4

ADC adjustmen t, 5-13

A9 pulse b oard, 3-97

p ow er atness adjustmen t, 5-14

ALC p ow er lev el accuracy, 3-67

step atten uator atness, 5-10

YO delay adjustmen t, 5-12

automated p erformance tests

AM/FM DA C o set and gain, 3-73

step atten uator atness, 5-5

AM input imp edance, 3-86

AM o set, 3-78

BASIC, 5-2

ampli er lter, 3-32

external trigger edge selection, 3-94

cop ying disks, 5-3

FM input imp edance, 3-89

soft w are installation, 5-3

fractional-N reference and API spurs, 12-5

A UX OUTPUT disconnection, 7-25

fractional-N V CO, 12-4

B

low p ow er SRD bias, 3-60

mo dulation generator atness, 3-92

BASIC

mo dulator o set and gain, 3-62

p ow er atness, 3-71

sampler assem bly, 3-10

selftest patc hes, 3-98

blo c king capacitor remo v al, 7-42

square w av e symmetry, 3-84

sw eep ramp, 3-9

remo v al, 7-25

BNC connectors, Option 004

YO driver +10 V reference, 3-14

remo v al, 7-36

YO driver gain and linearity, 3-15

YO lo op gain, 3-17

Agilent T ec hnologies oces, xi

calibration constan ts, 4-1

ALC p ow er lev el accuracy adjustmen t, 3-67

alphanumeric display remo v al, 7-10

AM accuracy adjustmen t, 3-75

AM delay adjustmen t, 3-80

AM dynamic range, 2-87

calibration constan ts utilit y, 5-20

Show 27 more labels / short text on this page
Index · 10 MHz standard · API1 spur, 12-5 · API2 spur, 12-5 · adjustmen ts · automated adjustmen ts · AM accuracy, 3-75 · AM delay, 3-80 · automated tests · binaries, 5-2 · FM gain, 3-82 · equipmen t, 5-2 · automated tests, 5-2 · binaries · automated tests, 5-2 · BNC connectors · v olts/GHz, 3-95 · c hanging, 4-6 · default, 4-7 · AM accuracy, 2-78 · de nition, 4-1 · descriptions, 4-8 · AM bandwidth, 2-84 · loading, 4-7 · memory areas, 4-2 · sa ving, 4-6 · Index-1
Page 604 of the Agilent Technologies 83640B user manual — P
Page 604 · click to enlarge
Page 604

c hec ksum v eri cation, 4-2

internal pulse accuracy, 2-105

coupler remo v al, 7-44

internal timebase aging rate, 2-7

D

J

DC blo c king capacitor remo v al, 7-42, 7-44

J10 remo v al, 7-25

dio de switc h remo v al, 7-44

J14 remo v al, 7-25

directional coupler remo v al, 7-44

J1 RF output connector remo v al, 7-13, 7-26

J1 RF output connector remo v al, front panel,

J1 RF output connector remo v al, Option 004,

RF dec k, 7-37

J6 remo v al, 7-25

J7 remo v al, 7-25

display lter remo v al, 7-17

K

display remo v al, 7-10

k eyb oard reassem bly, 7-18

doubler remo v al, 7-44

k eyb oard remo v al, 7-15

dual mo dulator remo v al, 7-44

E

line lter remo v al, 7-35

calibration constan ts, 4-2

lowband assem bly remo v al, 7-44

external ALC input remo v al, 7-36

low p ow er SRD bias adjustmen t, 3-60

external lev eling, 2-30

external trigger edge selection adjustmen t, 3-94

M

maxim um FM deviation, 2-100

F

maxim um lev eled p ow er, 2-28

micro circuit remo v al, 7-44

mo dulation generator atness adjustmen t, 3-92

mo dulation meter, 2-107

mo dulator o set and gain adjustmen t, 3-62

motherb oard disassem bly, 7-62

fan remo v al, 7-24

motherb oard remo v al, 7-74

O

lter remo v al, 7-44

op eration v eri cation, 2-4

op eration v eri cation form, 2-4

FM gain adjustmen t, 3-82

Option 004 BNC connectors

FM input imp edance adjustmen t, 3-89

remo v al, 7-36

fractional-N reference and API spurs adjustmen t,

P

fractional-N V CO adjustmen t, 12-4

frequency switc hing time, 2-18

front panel BNC connectors remo v al, 7-14

front panel disassem bly, 7-4

front panel em ulation, 5-17

front panel pro cessor remo v al, 7-9

front panel remo v al, 7-7, 7-65

line fuse replacemen t, 10-5

AM dynamic range, 2-87

external lev eling, 2-30

fuse housing remo v al, 7-34

Show 31 more labels / short text on this page
connector care, 10-7 · disassem bly · front panel, 7-4 · motherb oard, 7-62 · rear panel, 7-21 · display lter · cleaning, 10-4 · EEPR OM storage · line fuse, 10-5 · fan · noise, 10-2 · sp eed, 10-2 · fan lter · cleaning, 10-2 · lter · display, 10-4 · oces, xi · FM accuracy, 2-92 · FM bandwidth, 2-95 · passw ord, 4-3 · b ypassing, 4-5 · disabling, 4-3 · en tering, 4-3 · setting, 4-4 · p erformance tests · AM accuracy, 2-78 · fuse · AM bandwidth, 2-84 · rear panel, 10-5 · FM accuracy, 2-92 · Index-2
Page 605 of the Agilent Technologies 83640B user manual — T
Page 605 · click to enlarge
Page 605

rear panel, 7-27 , 7-70

frequency switc hing time, 2-18

rear panel interface, 7-31

internal pulse accuracy, 2-105

RF dec k, 7-73

RF output connector, 7-13, 7-26

maxim um FM deviation, 2-100

RPG, 7-12

maxim um lev eled p ow er, 2-28

source mo dule interface, 7-11

mo dulation meter, 2-107

p ow er accuracy, 2-23

v oltage selector switc h, 7-32

p ow er atness, 2-26

pulse mo dulation on/o ratio, 2-52

pulse mo dulation video feedthrough, 2-71

pulse p erformance, 2-55

pulse p erformance (alternate), 2-60

RF dec k, 7-37

single sideband phase noise, 2-48

spurious signals (harmonics & subharmonics),

RF dec k disassem bly, 7-37

spurious signals (line-related), 2-45

RF dec k remo v al, 7-73

spurious signals (non-harmonic), 2-41

RF output connector remo v al, 7-26

sw ept frequency accuracy, 2-11

RF output connector remo v al, front panel, 7-67

p ow er accuracy, 2-23

RF output connector remo v al, rear panel, 7-69

p ow er atness, 2-26

RPG1 rotary pulse generator remo v al, 7-12

p ow er atness adjustmen t, 3-71, 5-14

S

p ow er sensor calibration factor, 5-21

p ow er sensor con guration, 5-21

pre x n umb er, x

sampler assem bly adjustmen t, 3-10

pulse mo dulation on/o ratio, 2-52

selftest patc hes adjustmen t, 3-98

pulse mo dulation video feedthrough, 2-71

pulse p erformance, 2-55

serial n umb er, x

pulse p erformance (alternate test), 2-60

R

RF dec k, 7-39

RAM c hec ksum, 4-2

single sideband phase noise, 2-48

calibration constan ts, 4-2

rear panel BNC connectors remo v al, 7-25

source mo dule interface remo v al, 7-11

rear panel disassem bly, 7-21

spur adjustmen ts, 12-5

rear panel interface remo v al, 7-31

spurious signals (Harmonics & subharmonics),

rear panel remo v al, 7-27, 7-70

spurious signals (line-related), 2-45

k eyb oard, 7-18

spurious signals (non-harmonic), 2-41

square w av e symmetry adjustmen t, 3-84

10 MHz reference standard, 7-29

step atten uator atness

connectors, front panel BNC, 7-14

step atten uator atness p erformance test, 5-5

connectors, Option 004 BNC, 7-36

step atten uator remo v al, 7-44

connectors, rear panel BNC, 7-25

DC blo c king capacitor, 7-42

calibration constan ts, 4-2

sw eep ramp adjustmen t, 3-9

sw ept frequency accuracy, 2-11

front panel, 7-7 , 7-9, 7-65

T

front panel BNC connectors, 7-14

T1 transformer remo v al, 7-30

k eyb oard, 7-15

timebase oscillator adjustmen t, 3-4

transformer remo v al, 7-30

Show 27 more labels / short text on this page
FM bandwidth, 2-95 · internal timebase, 2-7 · transformer, 7-30 · replacemen t · front panel, 7-4 · motherb oard, 7-62 · rear panel, 7-21 · RF dec k · lifting up, 7-39 · sales oces, xi · self-test utility, 5-20 · service oces, xi · service p osition · RAM storage · soft w are · loading, 5-4 · reassem bly · remo v al · alphanumeric display, 7-10 · adjustmen t, 5-10 · storage · display lter, 7-17 · fan, 7-24 · fuse housing, 7-34 · line lter, 7-35 · motherb oard, 7-74 · Index-3
Page 606 of the Agilent Technologies 83640B user manual — U
Page 606 · click to enlarge
Page 606

U

Y

UVEPR OM

YIG oscillator remo v al, 7-44

calibration constan ts, 4-2

YO delay adjustmen t, 5-12

YO driver +10 V reference adjustmen t, 3-14

YO driver gain and linearity adjustmen t, 3-15

V CO tuning v oltage, 12-4

YO lo op gain adjustmen t, 3-17

v oltage selector switc h remo v al, 7-32

v olts/GHz adjustmen t, 3-95

Index-4

Ask about the Agilent Technologies 83640B

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Agilent Technologies 83640B

Be the first to ask — other owners or our team will answer.

Original document

Agilent Technologies 83640B manual PDF

606 pages·4.7 MB·Portable Generator·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…