
Agilent 81101A 50 MHz Pulse Generator
Reference Guide
S 1


Use the KNOB to select a mode or modify parameters
the required entry screen Press MORE key to access the additional screen menus: MODE/TRG TRG-LEV MEMCARD CONFIG Screen Selection Area

Agilent 81101A 50 MHz
Pulse Generator
Part No. 81101-91021 Printed in Germany March 2000
Edition 1.0, E0300

1998 Agilent Technologies 1998, 2000. All rights reserved.
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies Inc. as governed by United States and international copyright laws.
The material contained in this document is subject to change without notice. Agilent Technologies makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
This Agilent Technologies product has a warranty against defects in material and workmanship for a period of three years from date of shipment. During the warranty period, Agilent Technologies will, at its option, either repair or replace products that prove to be defective. For warranty service or repair, this product must be returned to a service facility designated by Agilent Technologies. The Buyer shall pay Agilent Technologies round-trip travel expenses. For products returned to Agilent Technologies for warranty service, the Buyer shall prepay shipping charges to Agilent Technologies and Agilent Technologies shall pay shipping charges to return the product to the Buyer. However, the Buyer shall pay all shipping charges, duties and taxes for products returned to Agilent Technologies from another country.
Copyright · Warranty

Agilent Technologies warrants that its software and firmware designated by Agilent Technologies for use with an instrument will execute its programming instructions when properly installed on that instrument. Agilent Technologies does not warrant that the operation of the instrument software, or firmware, will be uninterrupted or error free.
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by the Buyer, Buyer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance. No other warranty is expressed or implied. Agilent Technologies specifically disclaims the implied warranties of merchantability and fitness for a particular purpose.
The remedies supplied are the Buyer's sole and exclusive remedies. Agilent Technologies shall not be liable for any direct, indirect, special, incidental, or consequential damages, whether based on contract, tort or any other legal theory.
Agilent Technologies certifies that this product met its published specifications at the time of shipment. Agilent Technologies further certifies that its calibration measurements are traceable to the United States Institute of Standards and Technology, to the extent allowed by the Institute's calibrating facility, and to the calibration facilities of other International Standards Organization members.
Any adjustment, maintenance, or repair of this product must be performed by qualified personnel. Contact your customer engineer through your local Agilent Technologies Service Center. You can find a list of local service representatives on the Web at:
http://www.agilent.com/Service/English/index.html

The following general safety precautions must be observed during all phases of operation of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies Inc. assumes no liability for the customer's failure to comply with these requirements.
This product is a Safety Class 1 instrument (provided with a protective earth terminal). The protective features of this product may be impaired if it is used in a manner not specified in the operation instructions.
All Light Emitting Diodes (LEDs) used in this product are Class 1 LEDs as per IEC 60825-1.
This instrument is intended for indoor use in an installation category II, pollution degree 2 environment. It is designed to operate at a maximum relative humidity of 95% and at altitudes of up to 2000 meters. Refer to the specifications tables for the ac mains voltage requirements and ambient operating temperature range.
Verify that the product is set to match the available line voltage, the correct fuse is installed, and all safety precautions are taken. Note the instrument's external markings described under Safety Symbols on page 8.

To minimize shock hazard, the instrument chassis and cover must be connected to an electrical protective earth ground. The instrument must be connected to the ac power mains through a grounded power cable, with the ground wire firmly connected to an electrical ground (safety ground) at the power outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury.
Only fuses with the required rated current, voltage, and specified type (normal blow, time delay, etc.) should be used. Do not use repaired fuses or short-circuited fuse holders. To do so could cause a shock or fire hazard.
Do Not Operate in an Explosive Atmosphere
Do not operate the instrument in the presence of flammable gases or fumes.
Do Not Remove the Instrument Cover
Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be made only by qualified service personnel.
Instruments that appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel.
Safety Summary · Ground the Instrument · Fuses

Protective earth (ground) terminal
In the manuals:
Safety Summary · Safety Symbols

This guide provides reference information primarily for programming the Agilent 81101A via remote control.
Chapter 1 “General Programming Aspects” on page 13 gives general hints for programming instruments like the Agilent 81101A using SCPI commands.
Chapter 2 “Programming Reference” on page 25 provides detailed information on the SCPI commands supported by the instrument.
Chapter 3 “Specifications” on page 89 lists the instrument’s technical specifications and provides exact definitions for the instrument’s parameters.
For an introduction and information on the Agilent 81101A’s user interface, please refer to the Quick Start Guide, p/n 81101-91020.
About this Book

Conventions Used in this Book
This book uses certain conventions to indicate elements of the Agilent 81101A’s user interface. The following table shows some examples:
Softkeys Press the MODE/TRG softkey to access the Mode/ Trigger screen.
Hardkeys Press the MORE key to switch to the alternative softkey layout.
Alternate Keys Press SHIFT + 0 (ON/OFF) to switch on the output. The alternate key label—which is selected by pressing the SHIFT key—is given in parentheses.
Screen Quotes Move the entry focus down to PULSE-PERIOD and turn the knob to select INTERNAL PLL.
Entry Focus The highlight field, that can be moved with the cursor keys, to change modes, parameters, or parameter formats.
:VOLTage:HIGH 3V Full command for programming a 3 V high level. The upper case letters represent the short form of the command, which results in faster pro- gramming times.
*RST Common IEE 488 command, to reset instrument to default status.
About this Book



This chapter provides general information on writing GP-IB/SCPI programs for instruments like the Agilent 81101A.
Detailed information on programming the Agilent 81101A can be found in Chapter 2 “Programming Reference” on page 25.

General Programming Aspects The GP-IB Interface Bus
The General Purpose Interface Bus is the interface used for communication between a controller and an external device, such as the Agilent 81130A. The GPIB conforms to IEEE standard 488-1987, ANSI standard MC 1.1, and IEC recommendation 625-1.
If you are not familiar with the GPIB, please refer to the following books:
488.1-1987, IEEE Standard Digital Interface for Programmable Instrumentation.
488.2-1987, IEEE Standard Codes, Formats, and Common Commands for Use with IEEE Standard 488.1-1987.
In addition, the commands not from the IEEE 488.2 standard are defined according to the Standard Commands for Programmable Instruments (SCPI). For an introduction to SCPI and SCPI programming techniques, refer to the following document:
Instruments, published periodically by various publishers. To obtain a copy of this manual, contact the SCPI Consortium: SCPI Consortium Office Bode Enterprise 2515 Camino del Rio South, Suite 340 San Diego, CA, 92108 USA

General Programming Aspects Agilent 81101A Remote Control
GP-IB Address You can only set the GP-IB address from the front panel of the instrument (refer to the Quick Start Guide).
The default GP-IB address is 10.
The Agilent 81101A has two modes of operation:
The instrument is operated using the front panel keys.
After receiving the first command or query via the GP-IB, the instrument is put into remote state. The front panel is locked. To return to local operating mode, press SHIFT (LOCAL).
Modes of Operation

General Programming Aspects Programming Recommendations
Here are some recommendations for programming the instrument:
for setting the default setting is: *RST
programming speed. The device command for switching off the display is: :DISPlay OFF
For fast programming speed it is recommended to use the short forms. The short forms of the commands are represented by upper case letters. For example the short form of the command to set 100 ns double pulse delay is: :PULS:DOUB:DEL 100NS
subsystem command parts. Optional subsystem command parts are depicted in square brackets, e.g.: enable double pulse mode by [SOURce]:PULSe:DOUBle[1][:STATe] ON|OFF. Sufficient to use: :PULS:DOUB ON
period/frequency, you can explicitly specify output 1 (for compatibility reasons). If there is no output specified, the commands will set the default output 1. So, for setting a high level of 3 Volts for output 1 the commands are: :VOLT:HIGH 3V # sets high level of 3 V at out 1 :VOLT1:HIGH 3V # sets high level of 3 V at out 1
the instrument by setting it up manually. Enable the outputs so that the instrument’s error check system is on and possible parameter conflicts are immediately displayed.

General Programming Aspects Programming Recommendations
When you have found the correct setting, then use this to create the program. In the program it is recommended to send the command for enabling outputs (for example, :OUTPut ON) as the last command. With this procedure it is possible to switch off the error check system (:SYSTem:CHECk OFF) to increase programming speed. The error check is enabled again by sending *RST. *RST # set default settings :DISP OFF # switch off display update :SYST:CHEC OFF # switch off error check ... # other commands to set modes ... # and parameters :OUTP ON # enable the output
then send the common command *OPC?

General Programming Aspects Common Command Summary
This table summarizes the IEEE 488.2 common commands supported by the Agilent 81101A:
*CLS – Clear the status structure
*ESE Set the Standard Event Status register mask
*ESE? – Read the state of the Standard Event Status enable register
*ESR? – Read the state of the Standard Event Status event register
*IDN? – Read the Instrument's Identification string
*LRN? – Read the complete Instrument Setting
*OPC – Set the Operation Complete bit when all pending actions are complete
*OPC? – Read the status of the Operation Complete bit
*OPT? – Read the installed options
*RCL Recall a complete Instrument Setting from memory
*RST – Reset the instrument to standard settings
*SAV Save the complete Instrument Setting to memory
*SRE Set the Service Request Enable Mask
*SRE? – Read the Service Request Enable Mask
*STB? – Read the Status Byte
*TST? – Execute instrument’s selftest
*WAI – Wait until all pending actions are complete
Command Parameter Description · *TRG – Trigger


Voltage Warning Current Warning

0 1 2 3 4 5 6 7 8 9
0 1 2 3 4 5 6 7 8 9
Standard Event Status Operation Complete
Query Error Device Dependent Error
Execution Error Command Error
The instrument has a status reporting system conforming to IEEE 488.2 and SCPI. The above figure shows the status groups available in the instrument.
Each status group is made up of component registers, as shown in the following figure.
Timing Warning · Frequency Warning · Power On

General Programming Aspects Status Model
OR Hardware and Firmware condition
Condition Register
A condition register contains the current status of the hardware and firmware. It is continuously updated and is not latched or buffered. You can only read condition registers. If there is no command to read the condition register of a particular status group, then it is simply invisible to you.
Transition filters are used to detect changes of state in the condition register and set the corresponding bit in the event register. You can set transition filter bits to detect positive transitions (PTR), negative transitions (NTR) or both. Transition filters are therefore read/write registers. They are unaffected by *CLS.
An event register latches transition events from the condition register as specified by the transition filters or records status events. Querying (reading) the event register clears it, as does the *CLS command. There is no buffering, so while a bit is set, subsequent transition events are not recorded. Event registers are read only.

The enable register defines which bits in an event register are included in the logical OR into the summary bit. The enable register is logically ANDed with the event register and the resulting bits ORed into the summary bit. Enable registers are read/write, and are not affected by *CLS or querying.
Although all status groups have all of these registers, not all status groups actually use all of the registers. The following table summarizes the registers used in the instrument status groups.
CONDition NTR PTR EVENt ENABLe
QUEStionable √
OPERation1 x
Standard Event Status x x x √2 √3
Status Byte x x x √4 √5
1 Present, but not used. COND and EVEN always 0. 2 Use *ESR? to query. 3 Use *ESE to set, *ESE? to query 4 Use *STB? to query 5 Use *SRE to set, *SRE? to query

General Programming Aspects Status Model
The status byte summarizes the information from all other status groups. The summary bit for the status byte actually appears in bit 6 (RQS) of the status byte. When RQS is set it generates an SRQ interrupt to the controller indicating that at least one instrument on the bus requires attention. You can read the status byte using a serial poll or *STB?
0 Unused, always 0
1 Unused, always 0
2 Unused, always 0
3 QUESTionable Status Summary Bit
4 MAV—Message AVailable in output buffer
5 Standard Event Status summary bit
6 RQS; ReQuest Service
7 OPERation Status summary Bit, unused
0 Operation Complete, set by *OPC
1 Unused, always 0
3 Device Dependent Error
6 Unused, always 0

This Status Group is not used in the instrument.
0 Unused, always 0
1 Unused, always 0
2 Unused, always 0
3 Unused, always 0
4 Unused, always 0
5 Unused, always 0
6 Unused, always 0
7 Unused, always 0
8 Unused, always 0
9 Unused, always 0
10 Unused, always 0
11 Unused, always 0
12 Unused, always 0
13 Unused, always 0
14 Unused, always 0

General Programming Aspects Status Model
3 Unused, always 0
4 Unused, always 0
6 Unused, always 0
7 Unused, always 0
8 Unused, always 0
9 Unused, always 0
10 Unused, always 0
11 Unused, always 0
13 Unused, always 0
14 Unused, always 0
The QUEStionable Status group is used to report warning conditions amongst the voltage, current, pulse timing and frequency parameters. Warnings occur when a parameter, although not outside its maximum limits, could be causing an invalid signal at the output because of the actual settings and uncertainties of related parameters.

This chapter provides reference information on the following topics:
For general programming information, please refer to Chapter 1 “General Programming Aspects” on page 13.

Programming Reference Agilent 81101A SCPI Command Summary
:ARM (Trigger mode and source)
[:SEQuence[1] | :STARt]
[:LAYer[1]]
:EWIDTh
[:STATe] ON|OFF|1|0 Set/read External Width mode 39
:FREQuency Set/read trigger frequency, when PLL(INT2) used as source
:IMPedance Set/read impedance at EXT INPUT 40
:LEVel Set/read threshold level at EXT INPUT 40
:PERiod Set/read trigger period, when PLL(INT2) used as source
:SENSe EDGE|LEVel Set/read trigger on edge or gate on level 42
:SLOPe POS|NEG|EITH Set/read trigger slope at EXT INPUT 42
:SOURce IMM|INT2|EXT|MAN Set/read trigger source (VCO|PLL|EXT INPUT|MAN key)
:DISPlay
[:WINDow]
[:STATe] ON|OFF|1|0 Set/read front panel display state 43
:MMEMory
:CATalog? [A:] Read directory of memory card 44
:CDIRectory [] Change directory on memory card 44
:COPY [,A:],[,A:] Copy a file on memory card 45
:DELete [,A:] Delete a file from memory card 45
:INITialize [A:[DOS]] Initialize memory card to DOS format 45
:STATe , Load file from memory card to memory n 46
:STORe
:STATe , Store memory n to memory card 46
see page · Command Parameter Description

Programming Reference Agilent 81101A SCPI Command Summary
:OUTPut[1]
[:NORMal]
[:STATe] OFF|ON|1|0 Set/read normal output state
:IMPedance 47
[:INTernal] Set/read internal source impedance of output
:EXTernal Set/read expected external load imped- ance at output
:POLarity NORM|INV Set/read output polarity 48
[:SOURce]
:CURRent[1]
[:LEVel]
[:IMMediate]
[:AMPLitude] Set/read channel amplitude current 48
:OFFSet Set/read channel offset current 49
:HIGH Set/read channel high level current 50
:LOW Set/read channel low level current 51
:LIMit
[:HIGH] Set/read maximum current limits 52
:LOW Set/read minimum current limits 53
:STATe ON|OFF|1|0 Enable/Disable the current limits 53
:FREQency
[:CW | :FIXed] Set/read frequency of pulses 54
:AUTO ONCE Measure frequency at CLK-IN 55
:HOLD[1] VOLT|CURR Switch between VOLtage and CURRent command subtrees
:PHASe[1]
[:ADJust] Set/read channel phase 56
:PULSe
:DCYCLe[1] Set/read channel duty cycle 57
:DELay[1] Set/read channel delay (to leading edge) 57
:HOLD TIME|PRATio Hold absolute delay|phase delay fixed with varying frequency
:UNIT S|SEC|PCT|DEG|RAD Set/read delay units 59
see page · Command Parameter Description

Programming Reference Agilent 81101A SCPI Command Summary
:DOUBle[1]
[:STATe] OFF|ON Enable/disable double pulses per pulse period
:DELay Set/read delay between double pulses 60
:HOLD TIME|PRATio Hold absolute delay|phase delay fixed with varying frequency
:UNIT S|SEC|PCT Set/read delay units 61
:HOLD[1] WIDTh|DCYCle|TDELay Hold Width|Duty cycle|Trailing edge de- lay fixed with varying frequency
:PERiod Set/read pulse period 62
:AUTO ONCE Measure pulse period at CLK-IN 63
:TDELay[1] Set/read trailing edge delay 63
:TRANsition[1]
:HOLD TIME|WRATio Hold absolute transitions|transitions as width ratio fixed with varying width per period
:UNIT S|SEC|PCT Set/read transition time units 65
[:LEADing] Set/read leading edge transition 65
:TRAiling Set/read trailing edge transition 66
:AUTO OFF|ON|ONCE Couple trailing edge to leading edge 67
:TRIGger[1]
:VOLTage TTL|ECL Set/read TRIGGER|STROBE OUTput levels
:WIDTh[1] Set/read channel pulse width 68
[:SOURce]
:ROSCillator
:SOURce INTernal|EXTernal Set/read PLL reference source 68
:EXTernal
:FREQuency Set/read frequency of external PLL reference
:VOLTage[1]
[:LEVel]
[:IMMediate]
[:AMPlitude] Set/read channel amplitude voltage 70
see page · Command Parameter Description
![Page 29 of the Agilent Technologies 81101A user manual — [:ALL]](/cdn/reader/17/17a3a4734e4ad8ce5ca79410d0004b7554d8dbb00c25a104f8a3fe36779b7e0d/images/page-29.webp?w=400)
Programming Reference Agilent 81101A SCPI Command Summary
:OFFset Set/read channel offset voltage 71
:HIGH Set/read channel high level voltage 72
:LOW Set/read channel low level voltage 73
:LIMit
[:HIGH] Set/read maximum voltage limit 74
:LOW Set/read minimum voltage limit 74
:STATe ON|OFF|1|0 Enable|Disable the voltage limits 75
:STATus
:OPERation
[:EVENt]? Read Operation event register 75
:CONDition Numeric Read Operation condition register
:ENABle Numeric Set/Read Operation enable register
:NTRansition Numeric Set/Read Operation negative transition register
:PTRansition Numeric Set/Read positive transition register
:PREset Clear and preset status groups 76
:QUEStionable
[:EVENt]? Read Questionable event register 76
:CONDition? Read Questionable condition register
:ENABLe Numeric Set/Read Questionable enable register
:NTRansition Numeric Set/Read Questionable negative transi- tion register
:PTRansition Numeric Set/Read Questionable positive transi- tion register
:SYSTem
:CHECk
[:STATe] OFF Switch error checking off 78
:ERRor? Read error queue 78
:KEY Numeric Simulate key press or read last key pressed
:PRESet no function 81
:SECurity
see page · Command Parameter Description

Programming Reference Agilent 81101A SCPI Command Summary
[:STATe] ON|OFF Switch security on and off 82
:SET Block data Set/read complete instrument setting 83
:VERSion? Read SCPI compliance setting 83
:WARNing
[:COUNt]? Read number of active warnings 83
:STRing? Read active warnings as concatenated string
:BUFFer? Read maximum possible length of con- catenated string
:TRIGger
[:SEQuence [1] | :STARt] (Pulse mode and period source)
:COUNt Set/read number of triggered periods to be generated per ARM event
:IMPedance Set/read impedance at CLK-IN 85
:LEVel Set/read threshold level at CLK-IN 86
:SLOPe POS|NEG Set/read trigger slope at CLK-IN 86
:SOURce IMM | INT[1] | INT2 |EXT Set/read trigger source (IMM | VFO | PLL | CLK-IN)
see page · Command Parameter Description

Programming Reference Default Values, Standard Settings
Parameter *RST, Default Values
:ARM : EWIDth:STATe OFF
:FREQuency 100kHz
:IMPedance 50Ω
:LEVel +1.00V
:PERiod 10.00µs
:SENSe EDGE
:SLOPe POS
:SOURce IMMediate
:DISPlay [:WINDow] [:STATe] ON
:MMEMory :CATalog? not applicable
:CDIRectory not applicable
:COPY not applicable
:DELete not applicable
:INITialize not applicable
:LOAD :STATe not applicable
:STORe:STATe not applicable
:OUTPut[1][:NORMal][:STATe] OFF
:IMPedance[:INTernal] 50Ω
:EXTernal 50.0Ω
:POLarity NORMal
[:SOURce]:CURRent[1][:LEVel][:IMM][:AMPL] 20.0mA (from 50Ω into 50Ω)
:OFFset 0.0mA (from 50Ω into 50Ω)
:HIGH +10.0mA (from 50Ω into 50Ω)
:LOW –10.0mA (from 50Ω into 50Ω)
:LIMit[:HIGH] +10.0mA

Programming Reference Default Values, Standard Settings
Parameter *RST, Default Values
:LOW –10.0mA
:STATe OFF
:FREQ [:CW|:FIXed] 1.00MHz
:AUTO not applicable
:PHASe[1][:ADJust] 0.0
:PULSe:
:DCYCle[1] 10.0% (derived from Width and Period)
:DELay[1] 0.0
:UNIT S
:DOUBle[1][:STATe] OFF
:DELay 250 ns
:UNIT S
:HOLD[1] WIDTh
[:SOURce]:PULSe:PERiod 1µs
:AUTO not applicable
:TDELay[1] 100ns
:TRANsition[1|2]:HOLD TIME
[:LEADing] 5.0 ns
:TRAiling 5.0 ns
:TRIGger[1]:VOLTage TTL
:WIDTh[1] 100ns
:ROSCillator:SOURce INTernal
:EXTernal:FREQ 5MHz

Programming Reference Default Values, Standard Settings
Parameter *RST, Default Values
[:SOURce]:VOLTage[1]
:LEVel]
[:AMPLitude] 1.0V
:OFFSet 0.0V
:HIGH 500mV
:LOW –500mV
:LIMIt[:High] +500V
:STATe OFF
:STATus: :OPERation not applicable
:PRESet not applicable
:QUEStionable[:EVENt]? not applicable
:CONDition? not applicable
:ENABle not applicable
:NTRansition not applicable
:PTRansition not applicable
:SYSTem :CHECk [:ALL][:STATe] ON
:ERRor? not applicable
:PRESet not applicable
:SECurity[:STATe] OFF
:SET not applicable
:VERSion "1992.0
:WARNing[:COUNt]? not applicable
:STRing? not applicable
:BUFFer? not applicable
:TRIGger :COUNt 1
:IMPedance 50Ω
[IMMediate]

Programming Reference Default Values, Standard Settings
Parameter *RST, Default Values
:LEVel 1.0V
:SLOPe POSitive
:SOURce IMMediate

Programming Reference Programming the Instrument Trigger Modes
The following figure shows the instrument’s arming/triggering model:
:ARM Event detection layer
:ARM:SENSe EDGE (Triggered) LEVel (Gated) :ARM:SOURce IMMediate (Internal VFO) INTernal2 (Internal PLL) EXTernal (EXT INPUT) Manual (MAN key)
:TRIGger Event detection layer
:TRIGger:SOURce IMMediate (Internal VFO) INTernal2 (Internal PLL) EXTernal2 (CLK IN)
generate pulse period and increment counter
:TRIGger:COUNt n
COUNTER = :TRIG:COUNt?
You program the comprehensive triggering capabilities of the instrument using the SCPI :ARM and :TRIGger subsystems. Using these two command subsystems you can program the operating modes of the instrument, which are set up using the MODE/TRIGGER SCREEN on the front panel.

Programming Reference Programming the Instrument Trigger Modes
Use the :ARM subsystem to select the overall triggering mode of the instrument (CONTINUOUS, TRIGGERED, GATED, EXT WIDTH), and the :TRIGger subsystem to select the pulse period source, triggering and number of pulse periods per :ARM event (BURST or PATTERN length).
Set Continuous mode by arming the instrument from its internal oscillator:
:ARM:SOURce IMMediate Arm from internal oscillator.
Set Triggered mode by arming the instrument on edges from the EXT INPUT:
:ARM:SOURce EXTernal1 Arm from EXT INPUT :ARM:SENSe EDGE Arm on edge :ARM:SLOPe POSitive Arm on positive edge :ARM:LEVel 1V Set EXT INPUT threshold
You can also arm the instrument from the PLL and set the frequency (or period) of the PLL to the required triggering rate:
:ARM:SOURce INTernal2 Arm from PLL :ARM:SENSe EDGE Arm on edge :ARM:SLOPe POSitive Arm on positive edge :ARM:FREQuency Set PLL frequency
Set Gated mode by arming the instrument on levels from the EXT INPUT:
:ARM:SOURce EXTernal1 Arm from EXT INPUT :ARM:SENSe LEVel Arm on signal level :ARM:SLOPe POSitive Arm on positive level

Programming Reference Programming the Instrument Trigger Modes
Set External Width mode using the :EWIDth[:STATe] command:
:ARM:EWIDth ON Switch on EXT WIDTH mode
This command disables the arm-trigger system. The arm-trigger system is reenabled by switching OFF EWIDth mode.
Set Pulses mode by setting the :TRIGger:COUNt to 1 so that a single triggered pulse period is generated for every arm event. The trigger source sets the pulse period:
:TRIGger:COUNt 1 Single pulse period per arm event :TRIGger:SOURce INTernal1 Pulse period from internal osc. :DIGital:PATTern OFF Disable pattern data.
Pulse period Source :TRIGger SOURce
internal OSC. internal PLL CLK-IN
Set Burst mode by setting the :TRIGger:COUNt to the burst length required. The trigger source sets the pulse period for the pulses within the burst.
:TRIGger:COUNt 16 Burst of 16 pulse periods :TRIGger:SOURce INTernal1 Pulse period from internal osc. :DIGital:PATTern OFF Disable pattern data
INTernal[1] INTernal2 EXTernal2

The following reference sections list the instrument commands in alphabetical order. In addition to a command description, the attributes of each command are described under the following headings. Not all of these attributes are applicable to all commands. The commands are conform to the IEEE 488.2 SCPI standard.
Command Shows the short form of the command.
Long Shows the long form of the command.
Form Most commands can be used in different forms:
Set The command can be used to program the instrument
Query The command can be used to interrogate the instru- ment. Add a ? to the command if necessary.
Event The command performs a one-off action.
Parameter The type of parameter, if any, accepted by the command. The minimum and maximum value of numeric parameters can be accessed by the option MINimum or MAXimum.
Parameter Suffix The suffixes that may follow the parameter.
Any other commands that are implicitly executed by the command.
Value Coupling Any other parameter that is also changed by the command.
Range Coupling Any other parameters whose valid ranges may be changed by the command.
*RST value The value/state following a *RST command.
Specified Limits The specified limits of a parameter.
Absolute Limits Some parameters can be programmed beyond their specified limits.
Example Example programming statements.
Functional Coupling

Command :ARM:EWID
Long :ARM[:SEQuence[1]|STARt][:LAYer]:EWIDth[:STATe]
Parameter ON | OFF | 1 | 0
Description Use this command to enable the EXT WIDTH trigger mode available on the MODE/TRIGGER SCREEN. When EXT WIDTH mode is switched on, the rest of the :ARM and :TRIG system is disabled.
In EXT WIDTH mode a signal applied to the EXT INPUT determines the width and period of the output signal(s) from the instrument. You can still control the edge transition times and levels of the output signal(s).
Command :ARM:FREQ
Long :ARM[:SEQuence[1]|STARt][:LAYer]:FREQuency[:CW][:FIXed]
Parameter Suffix HZ with engineering prefixes, e.g.: MHZ is Megahertz.
*RST value 100 kHz
Specified Limits 1 mHz to 50 MHz
Description Use this command to program the frequency of the PLL (INTernal2) when it is used as the :ARM:SOURce for internal triggering of pulses, bursts or patterns.
If you are using the PLL as :TRIGger:SOURce to set the pulse frequency, use the [:SOURce]:FREQuency[:CW|:FIXed] command.
Example To set up bursts of four 20 MHz pulses occurring at a burst rate of 2 MHz:
:TRIG:SOUR INT :FREQ 20 MHZ :ARM:SOUR INT2 :ARM:SENS EDGE :ARM:FREQ 2 MHZ :TRIG:COUNT 4
Select internal osc. as pulse period source Set pulse frequency to 20 MHz Select PLL as triggering source Sense edge of PLL signal Set triggering frequency to 2 MHz Set burst length to 4
*RST value OFF · Parameter Numeric

Command :ARM:IMP
Long :ARM[:SEQuence[1]|STARt][:LAYer]:IMPedance
Parameter Suffix OHM with engineering prefixes, e.g.: MOHM is Megaohms.
*RST value 50 Ω
Specified Limits 50 Ω or 10 kΩ
Description Use this command to program the input impedance of the EXT INPUT connector. Note that only two settings are available. If you try to program any other value, it will be rounded to one of the specified values.
Example To set up the input impedance and the triggering threshold of the EXT INPUT connector:
Set EXT INPUT impedance to 50 Ω Set EXT INPUT threshold to 2.5 V
Command :ARM:LEV
Long :ARM[:SEQuence[1]|STARt][:LAYer]:LEVel
Parameter Suffix V with engineering prefixes.
*RST value +1.0 V
Specified Limits –10 V to +10 V
Description Use this command to program the triggering threshold of the EXT INPUT connector.
Example To set up the input impedance and the triggering threshold of the EXT INPUT connector:
Set EXT INPUT impedance to 50 Ω Set EXT INPUT threshold to 2.5 V
Parameter Numeric · Parameter Numeric

Command :ARM:PER
Long :ARM[:SEQuence[1]|STARt][:LAYer]:PERiod
Parameter Suffix S or SEC with engineering prefixes.
*RST value 10.00 µs
Specified Limits 20 ns to 999.5 s
Description Use this command to program the period of the PLL (INTernal2) when it is used as the :ARM:SOURce for internal triggering of pulses, bursts or patterns.
If you are using the PLL as :TRIGger:SOURce use the
[:SOURce]:PULSe:PERiod command to set the pulse period.
Example To set up bursts of four 50 ns pulses occurring every 500 ns:
:TRIG:SOUR INT :PER 50 NS :ARM:SOUR INT2 :ARM:SENS EDGE :ARM:PER 500ns :ARM:TRIG:COUNT 4
Select internal osc.as pulse period source Set pulse period to 50 ns Select PLL as triggering source Sense edge of PLL signal Set triggering period to 500 ns Set burst length to 4
Parameter Numeric

Command :ARM:SENS
Long :ARM[:SEQuence[1]|STARt][:LAYer]:SENSe
Parameter EDGE | LEVel
Description Use this command to select Triggered or Gated mode by choosing whether the instrument arms on the edge(s) or level of the arming signal.
When sensing edges, the instrument triggers when the arming signal crosses the selected threshold level (:ARM:LEV) in the selected direction (:ARM:SLOP). This corresponds to the Triggered mode selected on the Mode/Trigger screen when using the front panel.
When sensing levels, the instrument triggers as long as the arming signal is above (:ARM:SLOP POS), or below (:ARM:SLOP NEG) the selected threshold level (:ARM:LEV). This corresponds to the Gated mode selected on the MODE/TRIGGER SCREEN when using the front panel.
Command :ARM:SLOP
Long :ARM[:SEQuence[1]|STAR7t][:LAYer]:SLOPe
Parameter POSitive | NEGative | EITHer
Description Use this command to select the trigger slope for the arming signal when triggering on edges. Use EITHer to trigger on both the positive and negative edges of the arming signal. This allows you to trigger at twice the frequency of the arming signal.
If you are arming on levels, use this command to select whether the instrument triggers during the positive or negative cycle of the arming signal.
*RST value EDGE · *RST value POS

Command :ARM:SOUR
Long :ARM[:SEQuence[1]|STARt][:LAYer]:SOURce
Description Use this command to select the triggering mode of the instrument by selecting the source of the arming signal:
Triggering Source :ARM:SOURce Mode
Internal Osc. PLL EXT INPUT MAN key
IMMediate|INTernal[1] INTernal2 EXTernal1 MANual
Continuous Triggered | Gated by PLL Triggered | Gated by EXT IN Triggered | Gated by MANKey
Use :ARM:SENSe EDGE | LEVel to choose between Triggered and Gated.
Command :DISP
Long :DISPlay[:WINDow][:STATe]
Parameter ON | OFF | 1 | 0
Description Use this command to turn the front panel display on and off. Switching off the display improves the programming speed of the instrument.
*RST switches the display back on. Use :SYSTem:PRESet to perform an *RST without switching the display back on.
Example To switch off the front panel display:
Parameter IMMediate|INTernal[1]|INTernal2|EXTernal[1]|MANual · *RST value IMM · *RST value ON

Command :MMEM:CAT?
Long :MMEMory:CATalog?
Parameter ["A:"]
*RST value Not applicable
Description Use this command to get a listing of the contents of the currently selected directory on the memory card. As there is only one memory card slot, the parameter A: is optional. The information returned is: ,[,]
The total number of bytes used on the memory card. The total number of bytes still available on the memory card. String containing the name, type and size of one file: ",,"
The is always blank. A directory name has = 0.
Command :MMEM:CDIR
Long :MMEMory:CDIRectory
*RST value Not applicable
Description Use this command to change the current directory on the memory card. If you don't specify a directory name parameter, the root directory is selected.
Use the directory name ".." to move back to the parent directory of the current directory, unless you are already in the root directory "\".
Examples To change to different directories on the memory card:
Select root directory Select directory "PERFORM" Select parent directory

Command :MMEM:COPY
Long :MMEMory:COPY
Parameter "filename"[,"A:"],"copyname"[,"A:"]
Description Use this command to copy an existing file filename in the current directory to a new file copyname. If copyname is the name of a sub- directory in the current directory, a copy of the file filename is made in the sub-directory. Use ".." as copyname to copy a file into the parent directory of the current directory.
Examples To copy files on the memory card:
:MMEM:COPY "test1","test2" :MMEM:COPY "test1",".."
Copy test1 to test2 Copy test1 into parent directory
Command :MMEM:DEL
Long :MMEMory:DELete
Description Use this command to delete file filename from the currently selected directory.
Command :MMEM:INIT
Long :MMEMory:INITialize
Parameter ["A:"[,"DOS"]]
Description Use this command to initialize a memory card to DOS format.

Command :MMEM:LOAD:STAT
Long :MMEMory:LOAD:STATe
Parameter ,"filename"[,"A:"]
Specified Limits = 0 to 9 (integer)
Description Use this command to load a complete instrument setting from file filename in the current directory into memory in the instrument.
Memories 1 to 9 are the internal customer memories. Memory 0 holds the default setting.
Examples See next command.
Command :MMEM:STOR:STAT
Long :MMEMory:STORe:STATe
Parameter ,"filename"[,"A:"]
Specified Limits = 0 to 9 (integer)
Description Use this command to store a complete instrument setting from memory to file filename in the current directory on the memory card.
Memories 1 to 9 are the internal memories. Use memory 0 to store the current instrument setting to a file.
Examples To load/save instrument settings from/to the memory card:
Load FREQPERF into memory 1 Load AMPTEST as current setting Save current setting in memory 2 Recall memory 3 as current setting

Command :OUTP[1]
Long :OUTPut[1][:NORMal][:STATe]
Parameter ON | OFF | 1 | 0
Description Use this command to switch the normal OUTPUT on or off.
Example To switch on the output:
Command :OUTP[1]:IMP
Long :OUTPut[1]:IMPedance[:INTernal]
Parameter Suffix OHM with engineering prefixes, e.g.: MOHM is Megaohms.
*RST value 50 Ω
Specified Limits 50 Ω or 1 kΩ
Description Use this command to program the source impedance of the OUTPUT connector. Note that only two settings are available. If you try to program any other value, it will be rounded to one of the specified values.
Example To program the source impedance:
:OUTP:IMP 50OHM Set OUTPUT impedance to 50 Ω
Command :OUTP[1]:IMP:EXT
Long :OUTPut[1]:IMPedance:EXTernal
Parameter Suffix OHM with engineering prefixes, e.g.: MOHM is Megaohms.
*RST value 50.0 Ω
*RST value OFF · Parameter Numeric · Parameter Numeric

Specified Limits 0.1 Ω to 1 MΩ
Description Use this command to set the expected load impedance of the device under test at the OUTPUT connectors. If you have a non-50 Ω load, the output levels at the device under test will not be the levels you program or set via the front panel unless you set the expected load using this command.
Example To set the expected load impedance:
:OUTP:IMP:EXT 47.6OHM Set load impedance at OUTPUT to 47.6 Ω
Command :OUTP[1]:POL
Long :OUTPut[1]:POLarity
Parameter NORMal | INVerted
Description Use this command to invert the signal at the OUTPUT.
Example To invert and re-invert the signal:
Inverted signal at OUTPUT Normal signal at OUTPUT
Command :CURR[1]
Long [:SOURce]:CURRent[1][:LEVel][:IMMediate][:AMPLitude]
Parameter suffix A with engineering prefixes.
*RST value 20 mA (50 Ω into 50 Ω)
Specified Limits 10 V Outputs (from high Z into short): max. 400 mA typical
3.8V Outputs (50 Ω into short): max. 152 mA typical
*RST value NORM · Parameter Numeric

Amplitude = High – Low
Offset = High – Low
Description Use this command to program the amplitude current of the OUTPUT signal. Note that to set the OUTPUT levels in terms of current, you first have to execute the [:SOURce]:HOLD CURRent command to enable the [:SOURCE]:CURRENT subsystem.
The available current range is limited by the combination of:
:OUTPut:IMPedance:EXTernal
Example To program the amplitude current of the output signal:
Enable CURRENT subsystem Set OUTPUT amplitude to 75 mA
Command :CURR[1]:OFFSet
Long [:SOURce]:CURRent[1][:LEVel][:IMMediate]:OFFSet
Parameter suffix A with engineering prefixes.
*RST value 0.0 µA (50 Ω into 50 Ω)
Amplitude = High – Low
Offset = High – Low

Description Use this command to program the offset current of the OUTPUT signal. Note that to set the OUTPUT levels in terms of current, you first have to execute the [:SOURce]:HOLD CURRent command to enable the :SOURce]:CURRent subsystem.
The available current range is limited by the combination of:
Example To program the offset current of the output signal:
Enable CURRENT subsystem Set OUTPUT offset to 50 mA
Command :CURR[1]:HIGH
Long [:SOURce]:CURRent[1][:LEVel][:IMMediate]:HIGH
Parameter suffix A with engineering prefixes.
Amplitude = High – Low
Offset = High – Low
Range coupling Low level
*RST value +10 mA (50 Ω into 50 Ω)
Specified Limits 10 V Output (from high Z into short): –396 mA to 400 mA typical
3.8 V (from 50 Ω into short): –82 mA to 152 mA typical
Description Use this command to program the high level current of the OUTPUT signal. Note that to set the OUTPUT levels in terms of current, you first
Range coupling Amplitude · Parameter Numeric · Value coupling

have to execute [:SOURCE]:HOLD CURRent command to enable the [:SOURCE]:CURRent subsystem.
The available current range is limited by the combination of:
:OUTPut:IMPedance:EXTernal
Example To program the high level current of the output signal:
Enable CURRENT subsystem Set OUTPUT high level to 150 mA
Command :CURR[1]:LOW
Long [:SOURce]:CURRent[1][:LEVel][:IMMediate]:LOW
Parameter suffix A with engineering prefixes.
Amplitude = High – Low
Offset = High – Low
Range coupling High level
*RST value –10 mA (50 Ω into 50 Ω)
Specified Limits 10 V Outputs (from high Z into short): –400 mA to 396 mA typical
3.8 V Outputs (from 50 Ω into short): –84 mA to 150 mA typical
Description Use this command to program the low level current of the OUTPUT signal. Note that to set the OUTPUT levels in terms of current, you first have to execute the [:SOURce]:HOLD CURRent command to enable the [:SOURce]:CURRent subsystem.
The available current range is limited by the combination of:
Parameter Numeric · Value coupling

:OUTPUT:IMPedance:EXTernal
Example To program the low level current of the output signal:
Enable CURRENT subsystem Set OUTPUT low level to 50 mA
Command :CURR[1]:LIM
Long [:SOURce]:CURRent[1]:LIMit[:HIGH]
Parameter suffix A with engineering prefixes.
*RST value +10.0 mA
Description Use this command to set/read the high level current limit. If you switch on current limiting, the high level current cannot be set above the programmed limit.
The current is not limited by the OUTPUT hardware, this is a software limit.
Example To set the high level current limit for the output signal:
Enable CURRENT subsystem Set OUTPUT high level current limit to 50 mA Switch on OUTPUT limits
Parameter Numeric

Command :CURR[1]:LIM:LOW
Long [:SOURce]:CURRent[1]:LIMit:LOW
Parameter suffix A with engineering prefixes.
*RST value –10.0 mA
Description Use this command to set/read the low level current limit. If you switch on current limiting, the low level current cannot be set below the programmed limit.
The current is not limited by the OUTPUT hardware, this is a software limit.
Example To set the low level current limit for the output:
Enable CURRENT subsystem Set OUTPUT low level current limit to –50 mA Switch on OUTPUT limits
Command :CURR[1]:LIM:STAT
Long [:SOURce]:CURRent[1]:LIMit:STATe
Parameter ON | OFF | 1 | 0
Description Use this command to switch the output limits on or off. When you switch on the output limits, you cannot program the output levels beyond the programmed limits, until you switch off the output limits. The limits apply whether you program high/low levels or amplitude/offset levels.
Parameter Numeric · *RST value OFF

Example To set and activate the current limits for the output:
Enable CURRENT subsystem Set OUTPUT high level current limit to 50 m Set OUTPUT low level current limit to –50m Switch on OUTPUT limits
Command :FREQ
Long [:SOURce]:FREQuency[:CW|:FIXed]
Parameter Suffix Hz with engineering prefixes, or MHZ for Megahertz.
*RST value 1.00 MHz
Specified limits 1 mHz to 50 MHz
Description Use this command to set/read the pulse frequency. Select the frequency source for the pulse frequency using :TRIGger:SOURce. The currently selected source is programmed by this command. Note that the specified limits and available resolution depend on the selected source.
You cannot set the pulse frequency if you have selected the CLK-IN connector as the frequency source (:TRIG:SOUR EXT).
Example To set the pulse frequency to 40 MHz:
:TRIG:SOUR INT :FREQ 40MHz
Select internal osc. as pulse trigger Set pulse frequency to 40 MHz

Command :FREQ:AUTO
Long [:SOURce]:FREQuency[:CW|:FIXed]:AUTO
*RST value Not applicable
Description Use this command to measure the frequency at the CLK-IN connector. If the CLK-IN connector is the selected pulse frequency source, you can then read the measured value with :FREQ?
Example To measure and read the frequency at the CLK-IN connector:
Select ext CLK-IN as pulse trigger Measure frequency at CLK-IN Query pulse frequency
Command :HOLD
Long [:SOURce]:HOLD
Parameter VOLTage | CURRent
Description Use this command to enable either of the [:SOURce]:VOLTage or [:SOURce]:CURRent subsystems.
You can control the signal levels of the instrument’s OUTPUT in terms of voltage or current.
Form Event · Parameter ONCE · *RST value VOLT

Command :PHAS[1]
Long [:SOURce]:PHASe[1][:ADJust]
Parameter suffix DEG or RAD. A parameter without a suffix is interpreted as RAD.
Programming the pulse phase also executes [:SOURce]:PULSe:HOLD PHASe so that the pulse phase is held constant when the signal frequency is changed.
Delay = Phase 360 Period ×
*RST value 0.0
Specified limits 0 to 360° constrained by delay and period limits.
Description Use this command to set/read the relative phase delay of the output signal. This is equivalent to setting an absolute or percentage pulse delay with [:SOURce]:PULSe:DELay.
If you want the phase delay to remain constant when the pulse period is varied (rather than the absolute pulse delay) use [:SOURce]:PULSe:DELay[1]:HOLD PRATio.
Example To set and hold the phase delay:
Set OUTPUT phase to 180° Hold OUTPUT phase constant with varying period
Parameter Numeric · Functional coupling · Value coupling

Command :PULS:DCYC[1]
Long [:SOURce]:PULSe:DCYCle[1]
Width = Duty Cycle
*RST value 10.0% (derived from width and period)
Specified limits 0.001% to 99.9%, depends on width, transition & period.
Description Use this command to program the duty cycle of the pulse signal. If you want to set an absolute pulse width use [:SOURce]:PULSe:WIDTh[1].
If you want the pulse duty cycle to remain constant when the pulse period is varied (rather than the absolute pulse width use) :SOURce]:PULSe:HOLD[1] DCYCle
Example To set and hold the duty cycle:
Set OUTPUT duty cycle to 25% Hold duty cycle constant with varying period
Command :PULS:DEL[1]
Long [:SOURce]:PULSe:DELay[1]
Parameter suffix S with engineering prefixes. You can change the default unit using [:SOURce]:PULSe:DELay[1]:UNIT.
*RST value 0.0

Specified limits 0.00 ns to 999 s (limited by period – 20 ns)
Description Use this command to set/read the pulse delay. Delay is the time between the start of the pulse period and the start of the leading edge of the pulse.
If you want the pulse delay to remain constant when the pulse period is varied (rather than the phase delay) use [:SOURce]:PULSe:DELay[1]:HOLD TIME.
Example To set and hold the pulse delay:
Set OUTPUT delay to 500 ns Hold OUTPUT delay constant with varying period
Command :PULS:DEL[1]:HOLD
Long [:SOURce]:PULSe:DELay[1]:HOLD
Parameter TIME | PRATio
Description Use this command to set/read the coupling between the pulse period and the pulse delay:
TIME The absolute pulse delay is held fixed when the pulse period is varied (pulse phase varies).
PRATio The pulse phase delay (delay as ratio of period) is held fixed when the pulse period is varied (pulse delay varies).
Example To set and hold the pulse delay:
:PULS:DEL 500ns :PULS:DEL:HOLD TIME
Set OUTPUT delay to 500ns Hold OUTPUT delay constant with varying period
*RST value TIME

Command :PULS:DEL[1]:UNIT
Long [:SOURce]:PULSe:DELay[1]:UNIT
Parameter S | SEC | PCT | DEG | RAD
Description Use this command to set/read the default units for the pulse delay parameter. The default unit of a parameter is the unit used when the parameter is programmed to a value without a unit suffix.
Example To set the pulse delay to 50% of period:
Set OUTPUT delay unit to % Set OUTPUT delay to 50% of period
Command :PULS:DOUB[1]
Long [:SOURce]:PULSe:DOUBle[1][:STATe]
Parameter OFF | ON
Description Use this command to switch double-pulse mode on or off. In double- pulse mode two pulses are generated per pulse period. The delay between the leading edges of the first and second pulse can be adjusted.
*RST value S · *RST value OFF

Command :PULS:DOUB[1]:DEL
Long [:SOURce]:PULSe:DOUBle[1]:DELay
Parameter suffix S with engineering prefixes. You can change the default unit using [:SOURce]:PULSe:DOUBle:DELay[1]:UNIT.
*RST value 0.0
Specified limits 10 ns to 999.5 s (width +10 ns) to (period – width – 10 ns) min. period: 20 ns
Description Use this command to set/read the delay between the leading edges of the two pulses in double-pulse mode. The first pulse always starts at the start of the pulse period.
If you want the double delay to remain constant when the pulse period is varied (rather than the double delay as percentage of period) use [:SOURce]:PULSe:DOUBle[1]:DELay:HOLD TIME.
Example To set and hold the double-pulse delay:
Switch on Double pulses on OUTPUT Set interpulse delay to 500 ns Hold interpulse delay fixed with varying pulse period

Command :PULS:DOUB[1]:DEL:HOLD
Long [:SOURce]:PULSe:DOUBle[1]:DELay:HOLD
Description Use this command to set/read the coupling between the pulse period and the double-pulse delay:
TIME The absolute double-pulse delay is held fixed when the pulse period is varied.
PRATio The double-pulse delay as percentage of period is held fixed when the pulse period is varied.
Example To set and hold the double-pulse delay:
Switch on double-pulses on OUTPUT Set interpulse delay to 50% of pulse period Hold interpulse delay as fixed percent- age of pulse period
Command :PULS:DOUB[1]:DEL:UNIT
Long [:SOURce]:PULSe:DOUBle[1]:DELay:UNIT
Parameter S | SEC | PCT
Description Use this command to set/read the default units for the double delay parameter. The default unit of a parameter is the unit used when the parameter is programmed to a value without a unit suffix.
Example To set the double-pulse delay to 50%:
Set OUTPUT double delay unit to % Set OUTPUT double-pulse delay to 50% of period
Parameter TIME|PRATio · *RST value TIME · *RST value S

Command :PULS:HOLD[1]
Long [:SOURce]:PULSe:HOLD[1]
Parameter WIDTh | DCYCle | TDELay
Description Use this command to set whether the pulse width, the pulse duty cycle or the pulse trailing edge delay is held constant when the pulse period is changed.
Example To set and hold the delay and the duty cycle:
Hold OUTPUT delay fixed when frequency varies Set OUTPUT delay to 20ns Hold OUTPUT duty cycle fixed when frequen- cy varies Set OUTPUT duty cycle to 25%
:PULS:DCYC 25PCT
Command :PULS:PER
Long [:SOURce]:PULSe:PERiod
Parameter Suffix S with engineering prefixes.
Frequency = 1 Period
*RST value 1 µS
Specified limits 2 ns to 999.5 s
Description Use this command to set/read the pulse period. Select the pulse period source using :TRIGger:SOURce. The currently selected source is programmed by this command. Note that the specified limits and available resolution depend on the selected source.
*RST value WIDTh · Parameter Numeric · Value coupling

You cannot set the pulse period if you have selected the CLK-IN connector as the frequency source (:TRIG:SOUR EXT).
Example To set the pulse period using the internal oscillator:
Select internal osc. as pulse trigger Set pulse frequency to 25 ns
Command :PULS:PER:AUTO
Long [:SOURce]:PULSe:PERiod:AUTO
*RST value Not applicable
Description Use this command to measure the period at the CLK-IN connector. If the CLK-IN connector is the selected pulse period source, you can then read the measured value with :PULS:PER?
Example To measure the period at the CLK-IN connector:
Select ext. CLK-IN as pulse trigger Measure period at CLK-IN Query pulse period
Command :PULS:TDEL[1]
Long [:SOURce]:PULSe:TDELay[1]
Parameter Suffix S with engineering prefixes.
*RST value 100 ns
Specified Limits 10 ns to 999.5 s (max. period –10 ns)
Description Use this command to program the delay of the trailing edge of the pulse relative to the start of the pulse period. This is an alternative method of programming the pulse width.
Form Event · Parameter ONCE · Parameter Numeric

Example To program the pulse width by means of the delay parameters:
Set OUTPUT delay to 500 ns Hold OUTPUT delay constant with varying period Set OUTPUT trailing delay to 750 ns
Command :PULS:TRAN[1]:HOLD
Long [:SOURce]:PULSe:TRANsition[1]:HOLD
Parameter TIME | WRATio
Description Use this command to set the coupling between transition times and the pulse width:
TIME The absolute transition times are held when the pulse width is varied.
WRATio The ratio of transition time to pulse width is held when the pulse width is varied.
Example To set the coupling between transition times and the pulse width:
Hold OUTPUT transitions fixed when pulse width varies Hold OUTPUT transition width ratio when pulse width varies
*RST value TIME

Command :PULS:TRAN[1]:UNIT
Long [:SOURce]:PULSe:TRANsition[1]:UNIT
Parameter S | SEC | PCT
Description Use this command to set the default units for the pulse transition times. The default unit is used when the parameter is programmed to a value without a unit suffix.
Command :PULS:TRAN[1]
Long [:SOURce]:PULSe:TRANsition[1][:LEADing]
Parameter suffix S with engineering prefixes, or PCT
*RST value 5 ns
Specified limits 5 ns to 200 ms
By default: Trailing edge = Leading edge with:PULS:TRAN:TRA:AUTO ON.
Use :PULS:TRAN:TRA:AUTO OFF to enable independent programming of the trailing edge within a 1:20 ratio for the ranges.
Description Use this command to set/read the transition time of the pulse leading edge. Note that the leading and trailing edges of the pulse have to fit within the defined pulse width.
Example To set leading and trailing edges independently:
Set OUTPUT leading edge to 6 ns Enable independent setting of trailing edge Set OUTPUT trailing edge to 15 ns
*RST value S · Parameter Numeric · Parameter coupling

Command :PULS:TRAN[1]:TRA
Long [:SOURce]:PULSe:TRANsition[1]:TRAiling
Parameter suffix S with engineering prefixes, or PCT
*RST value 5 ns
Specified limits 5 ns to 200 ms
By default: Trailing edge = Leading edge with :PULS:TRAN:TRA:AUTO ON.
Use :PULS:TRAN:TRA:AUTO OFF to enable independent programming of the trailing edge within a 1:20 ratio for the ranges.
Description Use this command to set/read the transition time of the pulse trailing- edge. Note that the leading and trailing edges of the pulse have to fit within the defined pulse width.
Example To set the leading and trailing edges independently:
Set OUTPUT leading edge to 6ns Enable independent setting of trailing edge Set OUTPUT trailing edge to 15 ns
Parameter Numeric · Parameter coupling

Command :PULS:TRAN[1]:TRA:AUTO
Long :[SOURce]:PULSe:TRANsition[1]:TRAiling:AUTO
Description Use this command to set/read the automatic coupling of the pulse trailing edge transition time to the leading edge transition time.
ON The trailing edge transition time is automatically set to the same value as the leading edge, and is updated automatically each time the leading edge transition time changes.
OFF The trailing edge transition time is independently programmable.
ONCE The trailing edge transition time is set ONCE to the same value as the leading edge.
Example To set leading and trailing edges independently:
Set OUTPUT leading edge to 6 ns Enable independent setting of trailing edge Set OUTPUT trailing edge to 15 ns
Command :PULS:TRIG[1]:VOLT
Long [:SOURce]:PULSe:TRIGger[1]:VOLTage[:LEVel][:IMMediate] [:AMPlitude]
Parameter TTL | ECL
Description Use this command to set/read the output levels at the TRIGGER OUT connector.
Parameter ON|OFF|ONCE · *RST value ON · *RST value TTL

Command :PULS:WIDT[1]
Long [:SOURce]:PULSe:WIDTh[1]
Parameter suffix S with engineering prefixes
*RST value 100 ns
Specified limits 10 ns to 999.5 s (max. period –10 ns)
Description Use this command to program the width of the pulse signal. If you want to set width as duty cycle use [:SOURce]:PULSe:DCYCle[1].
If you want the pulse width to remain constant when the pulse period is varied (rather than the duty cycle) use [:SOURce]:PULSe:HOLD[1] WIDTh.
Example To set and hold the pulse width:
Set OUTPUT pulse width to 50 ns Hold pulse width constant with varying period
Command :ROSC:SOUR
Long [:SOURce]:ROSCillator:SOURce
Parameter INTernal | EXTernal
Description Use this command to set/read the reference source for the PLL. If you select the external reference (CLK-IN connector), you can choose to use a 5 MHz or 10 MHz reference signal using :ROSC:EXT:FREQ.
INTernal Lock the PLL to its internal reference
EXTernal Lock the PLL to a reference signal at the CLK-IN connector. The external refer- ence signal can be 5 or 10 MHz.
Parameter Numeric · *RST value INT

Example To set up the external PLL reference:
Set external PLL reference (CLK-IN) Set expected PLL reference frequency to 10 MHz
Command :ROSC:EXT:FREQ
Long [:SOURce]:ROSCillator:EXTernal:FREQuency
*RST value 5 MHz
Specified limits 5 MHz or 10 MHz
Description Use this command to set/read the expected reference frequency for the PLL at the CLK-IN connector. The external reference can be a 5 or 10 MHz signal. Note that if you program any value other than the two specified values, the value will be set to the nearest of the two specified values.
Example To set up the external PLL reference:
Set external PLL reference (CLK-IN) Set expected PLL reference frequency to 10 MHz
Parameter Numeric

Command :VOLT[1]
Long [:SOURce]:VOLTage[1][:LEVel][:IMMediate][:AMPLitude]
Parameter suffix V with engineering prefixes.
High = Offset + Amplitude
Low = Offset –
Range coupling With Offset, see page 71
*RST value 1.00 V
Specified limits 100 mVpp to 10.0 Vpp (values are valid from 50 Ω into 50 Ω)
Description Use this command to program the amplitude voltage of the output signal. Note that to set the output levels in terms of voltage, you first have to execute the [:SOURce]:HOLD VOLTage command to enable the [:SOURce]:VOLTage subsystem.
The available voltage range is limited by the combination of:
:OUTput:IMPedance:EXTernal
Example To set the amplitude voltage:
Enable VOLTAGE subsystem Set OUTPUT amplitude to 5 V
Parameter Numeric · Value coupling · 2 Amplitude

Command :VOLT[1]:OFFSet
Long [:SOURce]:VOLTage[1][:LEVel][:IMMediate]:OFFSet
Parameter suffix V with engineering prefixes.
High = Offset + Amplitude
Low = Offset – Amplitude
Range coupling With Amplitude, see page 70
*RST value 0.0 mV
Specified Limits –10 V to +10 V
Description Use this command to program the offset voltage of the OUTPUT signal. Note that to set the OUTPUT levels in terms of voltage, you first have to execute the [:SOURce]:HOLD VOLTage command to enable the [:SOURce]:VOLtage subsystem.
The available voltage range is limited by the combination of:
:OUTput:IMPedance:EXTernal
Example To set the offset voltage:
Enable VOLTAGE subsystem Set OUTPUT offset to –800mV
Parameter Numeric · Value coupling

Command :VOLT[1]:HIGH
Long [:SOURce]:VOLTage[1][:LEVel][:IMMediate]:HIGH
Parameter suffix V with engineering prefixes.
Amplitude = High – Low
Offset = High – Low
Range coupling With low level, see page 73
*RST value 500 mV
Specified limits –9.9 V to 10.0 V (50 Ω into 50 Ω)
Description Use this command to program the high level voltage of the OUTPUT signal. Note that to set the OUTPUT levels in terms of voltage, you first have to execute the [:SOURce]:HOLD VOLTage command to enable the [:SOURce]:VOLTage subsystem.
The available voltage range is limited by the combination of:
:OUTPut:IMPedance:EXTernal
Example To set the high level voltage:
Enable VOLTAGE subsystem Set OUTPUT high level voltage to 4.8 V
Parameter Numeric · Value coupling

Command :VOLT[1]:LOW
Long [:SOURce]:VOLTage[1][:LEVel][:IMMediate]:LOW
Parameter suffix V with engineering prefixes.
Amplitude = High – Low
Range coupling With high level, see page 72
*RST value –500 mV
Specified limits –10.0 V to 9.9 V (50 Ω into 50 Ω)
Description Use this command to program the low level voltage of the OUTPUT signal. Note that to set the OUTPUT levels in terms of voltage, you first have to execute the [:SOURce]:HOLD VOLTage command to enable the [:SOURce]:VOLTage subsystem.
The available voltage range is limited by the combination of:
:OUTPut:IMPedance:EXTernal
Example To set the low level voltage:
Enable VOLTAGE subsystem Set OUTPUT low level to 500mV

Command :VOLT[1]:LIM
Long [:SOURce]:VOLTage[1]:LIMit[:HIGH]
Parameter suffix V with engineering prefixes.
*RST value +500 mV
Description Use this command to set/read the high level voltage limit. If you switch on voltage limiting, the high level voltage cannot be set above the programmed limit. Note that the voltage is not limited by the OUTPUT hardware, this is a software limit.
Example To set and activate the high level voltage limit:
Enable VOLTAGE subsystem Set OUTPUT high level limit to 3 V Switch on OUTPUT limits
Command :VOLT[1]:LIM:LOW
Long [:SOURce]:VOLTage[1]:LIMit:LOW
Parameter suffix V with engineering prefixes.
*RST value –500 mV
Description Use this command to set/read the low level voltage limit. If you switch on voltage limiting, the low level voltage cannot be set below the programmed limit. Note that the voltage is not limited by the OUTPUT hardware, this is a software limit.
Example To set and activate the low level limit:
Enable VOLTAGE subsystem Set OUTPUT low level voltage Switch on OUTPUT limits
Parameter Numeric · Parameter Numeric

Command :VOLT[1]:LIM:STAT
Long [:SOURce]:VOLTage[1]:LIMit:STATe
Parameter ON | OFF | 1 | 0
Description Use this command to switch the output limits on or off. When you switch on the output limits, you cannot program the output levels beyond the programmed limits, until you switch off the voltage limits. The limits apply whether you program high/low levels or amplitude/offset levels.
Example To set and activate the high and low voltage limits:
Enable VOLTAGE subsystem Set OUTPUT high level voltage limit to 3 V Set OUTPUT low level voltage limit to 0 V Switch on OUTPUT limits
Command :STATus:OPERation
This command tree accesses the OPERation status group. The OPERation status group is not used by the instrument. Therefore, this command tree is redundant.
*RST value OFF

Command :STATus:PRESet
Long :STATus:PRESet
*RST value Not Applicable
Status Group Register Preset value
NTR 0000000000000000
PTR 0111111111111111
NTR 0000000000000000
Command :STATus:QUEStionable
This command tree accesses the QUEStionable status group. The QUEStionable status group contains warning bits for voltage, current, time and frequency parameters. A warning occurs when the output signal could be out of specification due to the combined specification uncertainties of many parameters, although all parameters are set within their individually specified limits. If a parameter is set outside its specified limits an error is generated.
The following commands are used to access the registers within the status group:

Form *RST value Description
Query Not Applicable This command reads the event register in the QUEStionable status group.
Form *RST value Description
Query Not Applicable This command reads the condition register in the QUEStionable status group.
Form Parameter *RST value Specified limits Description
Set & Query Numeric Not affected by *RST'' 0 – 32767
This command sets or queries the enable register in the QUEStionable status group.
Form Parameter *RST value Specified limits Description
Set & Query Numeric Not applicable' 0 – 32767
This command sets or queries the negative transition register in the QUEStionable status group.
Form Parameter *RST value Specified limits Description
Set & Query Numeric Not applicable' 0 – 32767
This command sets or queries the positive transition register in the QUEStionable status group.

Command :SYST:CHEC
Long :SYSTem:CHECk[:ALL][:STATe]
Parameter OFF | ON
Description Use this command to switch the instrument’s error checking on or off. Switch off the error checking if you want to improve the programming speed of the instrument, but remember that no invalid parameter or mode settings will be detected and reported. Error checking is switched off by the *RST command, or when default setting is invoked.
Command :SYST:ERR?
Long :SYSTem:ERRor?
*RST value Not Applicable
Description Use this command to read the instrument error queue. The instrument error queue can store up to 30 error codes on a first-in-first-out basis. When you read the error queue, the error number and associated message are put into the instrument’s output buffer.
If the queue is empty, the value 0 is returned, meaning NO ERROR. If the queue overflows at any time, the last error code is discarded and replaced with –350 meaning QUEUE OVERFLOW.
Example To read the error queue:
:SYS:ERR? Query for errors
–222 "Data out of range" overlap at output 1: Width>Double Delay
*RST value ON · Form Query · Example Error String

The above message is an example of a customized description. Generic descriptions are available in the SCPI 1995 Command Reference, items 21.8.4 to 21.8.11.
For more detailed information in the 81110A error.
Send ":SYST:WARN:STR?". Alternatively, the HELP key shows the current errors and warnings and their description on the instruments display.
Command :SYST:KEY
Long :SYSTem:KEY
Parameter suffix No suffix allowed
No. Key Description
–1 No key pressed (Query only)
1 DATA ENTRY 1
2 DATA ENTRY 2
3 DATA ENTRY 3
4 DATA ENTRY 4
5 DATA ENTRY 5
6 DATA ENTRY 6
7 DATA ENTRY 7
8 DATA ENTRY 8
9 DATA ENTRY 9
10 DATA ENTRY .
Parameter Numeric · *RST value –1 · Specified limit

No. Key Description
16 MAN
17 STORE
18 HELP
19 SHIFT
20 MORE
25 NANO
26 MICRO /M EGA
27 MILLI / KILO
28 ENTER
29 Modify Knob Left (counter-clockwise)
30 Modify Knob Right (clockwise)
Description In query form, this command reads the last key pressed. The buffer is emptied by *RST and returns the value –1 when empty.
In set form, the command simulates pressing a key on the front panel. Simulated key press are also recorded as the last key pressed.
:SYST:KEY 19 sets the instrument to LOCAL mode.

(LOCAL) key are active. Because the instrument normally switches to remote mode when any command is received, including :SYSTem:KEY, simulating one of the other disabled keys has no effect.
the instrument from entering remote mode by using the REN line of the GP-IB to maintain local mode (LOCAL 7 in BASIC).
If you do this, the :SYSTem:KEY command is the only command that works. Any other commands will be buffered in the instrument blocking any further :SYSTem:KEY commands, until remote mode is enable.
Command :SYST:PRES
Long :SYSTem:PRESet
Form No function.

Command :SYST:SEC
Long :SYSTem:SECurity[:STATe]
If you accidentally switch on system security, and want to rescue the settings stored in the instrument, store the settings on a memory card.
You can then recall them from the memory card later.
Use this command to switch on system security mode. Switch on system security if you need to make sure that all instrument settings stored in the instrument are erased automatically when the instrument is switched off, or when security mode is switched off.
The instrument settings are erased by overwriting them with the default settings.
System security mode is not available via the front panel. If you want to erase all settings by hand:
1 SHIFT + STORE (RECALL) + 0 to recall the default settings from memory location 0.
2 STORE + 1, STORE + 2, ... STORE + 9, to store the defaults in memory locations 1 to 9.
Parameter ON|OFF · *RST value OFF · Description

Command :SYST:SET
Long :SYSTem:SET
*RST value Not applicable
Description In query form, the command reads a block of data containing the instrument’s complete setup. The set-up information includes all parameter and mode settings, but does not include the contents of the instrument setting memories, the status group registers or the :DISPlay[:WINDow][:STATe] The data is in a binary format, not ASCII, and cannot be edited.
In set form, the block data must be a complete instrument setup read using the query form of the command.
Command :SYST:VERS?
Long :SYSTem:VERSion?
*RST value "1992.0"
Description Use this command to read the SCPI revision to which the instrument complies.
Command :SYST:WARN?
Long :SYSTem:WARNing[:COUNt]?
*RST value Not applicable
Description Use this command to read the number of warnings which are currently active. Note that the warning status of voltage, current, time and frequency are also summarized by bits in the QUESTionable Status register.
Parameter Block data · Form Query · Form Query

Command :SYST:WARN:STR?
Long :SYSTem:WARNing:STRing?
*RST value Not applicable
Description Use this command to read all the currently active warning messages. The warning messages are concatenated to form a single string with a “;” as separator between the messages.
Command :SYST:WARN:BUFF?
Long :SYSTem:WARNing:BUFFer?
*RST value Not applicable
Description Use this command to read the maximum possible number of characters that could be returned by :SYST:WARN:STR? if all warnings were active.
Command :TRIG:COUNt
Long :TRIGger[:SEQuence[1]]:COUNt
Specified limits 1 to 65 536
Description Use this command to set/read the number of trigger events (pulse periods) to be generated for each arming event. This corresponds to selecting the event mode on the MODE/TRIGGER SCREEN:
PULSES Set a trigger count to 1 so that a single pulse period is generated for each arming event. The instrument is in pulse (stream) mode.
BURST of Set a trigger count of 2 to 65536 so that a burst of 2 to 65536 pulse periods is generated for each arming event.

Examples To set up a triggered burst of 16 Single Pulses at Out1, each burst triggered by a positive edge at the EXT INPUT:
:ARM:SOUR EXT1 :ARM:SENS EDGE :ARM:SLOP POS :TRIG:COUN 16 :TRIG:SOUR INT1 :DIG:PATT OFF :PULS:DOUB OFF
Set arming from EXT INPUT Set arming on edges Set arming on positive edges Burst length 16 Pulse period trigger from internal osc. Disable pattern operating mode Ensure single pulses at OUTPUT
To set up gated pulses (Single Pulses at Out1), gated by a positive level at the EXT INPUT:
:ARM:SOUR EXT1 :ARM:SENS LEV :ARM:SLOP POS :TRIG:COUN 1 :TRIG:SOUR INT1 :DIG:PATT OFF :PULS:DOUB OFF
Set arming from EXT INPUT Set arming on levels Set arming on positive level 1 pulse period Single pulse output mode Pulse period trigger from internal osc. Disable pattern data Ensure single pulses at OUTPUT
Command :TRIG:IMP
Long :TRIGger:IMPedance
Parameter Suffix OHM with engineering prefixes, e.g.: MOHM is Megaohms.
*RST value 50 Ω
Specified Limits 50 Ω or 10 kΩ
Description Use this command to program the input impedance of the CLK-IN connector. Note that only two settings are available. If you try to program any other value, it will be rounded to one of the specified values.
Example To set the input impedance and the threshold of the CLK-IN connector:
Set CLK-IN impedance to 50 Ω Set CLK-IN threshold to 2.5V Pulse period trigger from CLK-IN
Parameter Numeric

Command :TRIG:LEV
Long :TRIGger:LEVel
Parameter Suffix V with engineering prefixes.
*RST value 1.0 V
Specified Limits –10 V to +10 V
Description Use this command to program the triggering threshold of the CLK-IN connector.
Example To set the input impedance and the threshold of the CLK-IN connector:
Set CLK-IN impedance to 50 Ω Set CLK-IN threshold to 2.5V Pulse period trigger from CLK-IN
Command :TRIG:SLOP
Long :TRIGger:SLOPe
Parameter POSitive | NEGative
Description Use this command to select the trigger slope for the pulse period triggering signal applied to the CLK-IN connector.
Command :TRIG:SOUR
Long :TRIGger:SOURce
Parameter IMMediate | INTernal[1] | INTernal2 | EXTernal2
Description Use this command to select the pulse period source of the Agilent 81101A by selecting the source of the pulse period trigger signal:
Parameter Numeric · *RST value POS · *RST value IMM

Pulse period sources set by :TRIG:SOUR
Pulse period source:TRIG:SOURce
internal osc internal PLL CLK-IN
IMMediate INTernal[1] INTernal2 EXTernal2



In this chapter you will find the specifications of the Agilent 81101A Pulse Generator.
At the end of this chapter, “Pulse Parameter Definitions” on page 105 provides detailed information on the definition of the pulse parameters used by the instrument.
Specifications describe the instrument’s warranted performance. Non- warranted values are described as typical. All specifications apply after a 30 minute warm-up phase with 50 Ohm source, a 50 Ohm load resistance and separate channels. They are valid from 0 °C to 55 °C ambient temperature.

Specifications Declaration of Conformity
Manufacturer Agilent Technologies Deutschland GmbH Boeblingen Verification Solutions Herrenberger Str. 130
We declare that the system:
AGILENT 81100 Family of Pulse-/Data Generators AGILENT 81110 A 330/165 MHz Pulse/Pattern Generator AGILENT 81104 A 80 MHz Pulse Pattern Generator AGILENT 81101 A 50 MHz Pulse Pattern Generator AGILENT 81112 A 330 MHz , 3.5V Output Module- AGILENT 81130 A * 400/660 MHz Puls-/Pattern Generator AGILENT 81131 A * 400 MHz , 3.5V Output Module AGILENT 81132 A * 660 MHz , 2.5V Output Module AGILENT E 8305 A * VXI Plugin 250 MHz Pulse Generator AGILENT E 8306 A * VXI Plugin 100 MHz Clock Generator
Conforms to the following standards:
Safety IEC 1010-1:1990 +A1:1992 +A2 :1995 ... EN61010-1:1993 +A2:1995
EMC EN 55011:1991 / CISPR 11 Group 1, Class B* EN 55011:1991 / CISPR 11 Group 1, Class A EN 61000-4-2:1995 ESD: 4kVcd; 8 kVad;4kV c.p. EN 61000-4-3:1995 Radiated Immunity: 3V/m 80%AM ENV 50204: 1995 Radiated Immunity: 3V/m;50%Dty EN 61000-4-4:1995 Fast Transients/Bursts: 0.5kV, 1kV EN 61000-4-5:1995 Surges: 1kVdiff; 2kV com.mode EN 61000-4-6:1995 Conducted Immunity EN 61000-4-8:1993 Power freq. magn. field 3A/m;50Hz IEC1000-4-11:1994 Voltage Dips and Interruptions
The product herewith complies with the requirements of the • Low Voltage Directive (73/23/EEC)
During the measurements against EN55011, the I/O ports were terminated with their nominal impendance, the GP-IB connection was terminated with the cable AGILENT 10833B.
When the Product is connected to other devices, the user must ensure that the connecting cables and the other devices are adequately shielded to prevent radiation.
Böblingen, June 09th 1998 Wolfgang Fenske Update, Oct. 13th 1998 Regulations Consultant
71034 Böblingen/Germany · Supplementary Information

Specifications Agilent 81101A Specifications
IEC1010, CSA1010
Power requirements
100–240 Vac, ±10%, 50–60 Hz;
100–120 Vac, ±10%, 400 Hz
Power consumption: 300 VA max.
89 mm x 426 mm x 521 mm

Specifications Agilent 81101A Specifications
8.5 kg Single Channel 9.2 kg Dual Channel
13.8 kg Dual Channel
Warranty
Acoustic Noise Emission
For ambient temperature up to 30°C, under normal operation and at the typical operator position:
LpA = 52 dB (5.9 bel) typical {47 dB (5.3 bel) at 23°C) typical}
Measured in accordance with ISO 7779/EN 27779.

Specifications Agilent 81101A Specifications
Period can also be entered as frequency.
Resolution: 3.5 digits, 5 ps best case for VFO 4 digits, 1 ps best case for PLL
There are 2 period generation sources available:
You can sweep your timing values without danger of spurious pulses or drop outs that could cause measurement errors. This applies to continuous mode with timing values < 100 ms (frequency: < 10 Hz), and consecutive values between one-half and twice the previous value.
Period Agilent 81101A

Specifications Agilent 81101A Specifications
Can be entered as absolute width, duty cycle or trailing edge delay.
Duty cycle values from 0.1% to 95% can be entered directly. For values >95 % press SHIFT and use the Modify knob. Note that pulses may be deteriorated or skipped due to the inaccuracy of period and width. Hence for large values, it is better to select complement and enter 100 minus the required duty cycle value.
Measured between trigger output and main output. Can be entered as absolute delay, phase ° or % of period.
Fixed delay from TRIGGER OUT:
17.0 ns typical
Width Agilent 81101A · Delay Agilent 81101A

Specifications Agilent 81101A Specifications
Double pulse delay and delay are mutually exclusive. Double Pulse delay is the delay between the two pulses in Double Pulse mode.
Double Pulse Delay Agilent 81101A
Double Pulse Delay range:
20 ns to 999.5 s (width + 10 ns) to (period – width – 10 ns)
Measured between 10% and 90% of amplitude. Can be entered as leading/ trailing edge or % of width.
Transition Times Agilent 81101A
7.5 ns typical for 1 kΩ source impedance

Specifications Agilent 81101A Specifications
Leading and trailing edges can be programmed independently within the following ranges (Maximum ratio 1:20):

Level Parameters Agilent 81101A
ON/OFF: relay connect/disconnect output (HiZ).
For loads ≠ 50Ω, the actual load impedance can be entered to correct the output values.
5ns

Specifications Agilent 81101A Specifications
Level parameters can be entered as high/low level in terms of voltage or current or offset/amplitude.
Level Specifications (50Ω into 50Ω) (1kΩ into 50Ω)
Level Window –10.0 V to +10.0 V –20.0 V to + 20.0 V
Level Accuracy (in + 19 V level window):
+(3% Amplitude + 75 mV) +(3% Amplitude + 150 mV)
Pulse Performance Agilent 81101A
Dynamic Crosstalk < 0.1% typical
It is possible to select between three clock sources, the startable oscillator (VFO), the internal PLL, or the External Clock. In Triggered Mode the PLL can be used as the trigger source for the VFO, without the need of an additional source.

Specifications Agilent 81101A Specifications
Input Specifications Agilent 81101A
Delay from Clock Input to TRIGGER OUT/ STROBE OUT:
Rear panel BNC connector used as:
The input frequency can be measured.
Clk In (5 MHz or 10 MHz selectable) or to its internal reference.
When locked to the internal reference, period accuracy, resolution, and jitter are improved. When locked to an external frequency reference, the external frequency affects these accuracies.
trigger source, while the output period is determined by the startable oscillator.
12 ns typical

Specifications Agilent 81101A Specifications
Frequency accuracy can be increased by using a precise external clock.
synchronous to the signal at clock input. The signal at the External Input is used for arming.
The output signal consists of single or double pulses, controlled by the Trigger mode.
The output signal consists of bursts of single or double pulses, controlled by the Trigger mode.
Burst Parameters Agilent 81101A

Specifications Agilent 81101A Specifications
Generate continuous pulses, double pulses, or bursts.
Each active input transition (rising, falling or both) triggers a single pulse, a double pulse, or a burst.
The trigger source can be selected from:
The active input level (high or low) enables pulses, double pulses, or bursts. The last pulse, double pulse, or burst is always completed. The gate source can be selected from:
To recover a pulse shape of an external signal, applied to the External Input, the period and width are maintained, levels and transitions can be set.

Specifications Agilent 81101A Specifications
Trigger Output Specifications Agilent 81101A
Trigger pulse width: typically 50% of period Triggered mode: 9 ns typically External Width mode: recovered pulse shape of external signal
Output Voltage TTL or ECL into GND selectable
Strobe Output Specifications Agilent 81101A
Output Voltage TTL or ECL into GND selectable

Specifications Agilent 81101A Specifications
typ. value
External Width Ext Input Strobe/Trigger Out OUTPUT
8.5 ns 22.5 ns
Trigger Gated Ext Input Strobe/Trigger Out OUTPUT
12.0 ns 29.0 ns
Continuous Strobe/ Trigger Out
OUTPUT 17.0 ns
Ext. clock signal as pulse period
CLK-IN Strobe/Trigger Out OUTPUT
12.0 ns 29.0 ns
Mode from to

Specifications Agilent 81101A Specifications
Parameter values can be entered exceeding the specified range.
Error messages indicate conflicting parameters.
Displays a context-sensitive message about the selected parameter. Concept help for getting started is also available. If warnings or errors occur, the HELP key displays the warning/error list accordingly.
Actual setting is saved on power down. 9 user and 1 default setting are also stored in instrument.
99 settings can be stored per 1 MB (MS-DOS, PCMCIA) memory card. Also used for convenient firmware updates.

Specifications Agilent 81101A Specifications
Operates according to IEEE standard 488.2, 1987 and SCPI 1992.0.
SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT1,C0.
(all checks and display off).
Command Typical execution time
One parameter or mode 30 ms typ.
Recall Setting 250 ms typ


Here you find the pulse parameter definitions of terms used in the instrument specifications. In the following figure a graphical overview of the pulse parameters is provided:


Specifications Pulse Parameter Definitions
The time reference point is at the median of the amplitude (50% amplitude point on pulse edge):
The time interval between the leading edge medians of consecutive output pulses:
Interval between trigger point of the external trigger input signal and the trigger output pulse’s leading edge median.
Interval between leading and trailing edge medians:
The specified and displayed value is that obtained with fastest edges, essentially equal to the interval from the start of the leading edge to the start of the trailing edge. By designing so that the pulse edges turn about their start points, the interval from leading edge start stays unchanged (in

Specifications Pulse Parameter Definitions
practice, start points may shift with changes in transition time) when transition times are varied. This is more convenient for programming and the width display is easy to interpret.
Interval between leading edge medians of trigger output pulse and output pulse:

Variable Delay Fixed Delay
The specified and displayed value is that obtained with the fastest leading edge. Pulse delay has two components, a fixed delay from trigger output to output signal and a variable delay with respect to the trigger output.
Interval between leading edge medians of the double pulses.

Interval between corresponding leading edge medians of the output signals.
Trigger Output · Output Signal

Specifications Pulse Parameter Definitions
Interval between the 10% and 90% amplitude points on the leading/ trailing edge:

Peak deviation of an edge from a straight line through the 10% and 90% amplitude points, expressed as percentage of pulse amplitude:

0% · 100 % Amplitude · 0 % Amplitude

Specifications Pulse Parameter Definitions
Short-term instability of one edge relative to a reference edge. Usually specified as rms value, which is one standard deviation or “sigma”. If distribution is assumed Gaussian, six sigma represents 99.74% of the peak-peak jitter.
The reference edge for period jitter is the previous leading edge. That for delay jitter is the leading edge of the trigger output. Width jitter is the stability of the trailing edge with regard to the leading edge.
Long-term average instability over a specific time, for example, hour, year. Jitter is excluded.
Pulse output is specified as pulse top and pulse base (usually referred to as high level and low level), or as peak to peak amplitude and median offset. A “window” specification shows the limits within which the pulse can be positioned.

Max High Level · Low Level · Min Low Level

Specifications Pulse Parameter Definitions
Preshoot and overshoot are peak distortions preceding/following an edge. Ringing is the positive-peak and negative-peak distortion, excluding overshoot, on pulse top or base. For example, a combined preshoot, overshoot, and ringing specification of 5% implies:

Time taken for pulse levels to settle within level specifications, measured from 90% point on leading edge.

Overshoot eg 5%

Specifications Pulse Parameter Definitions
When an instrument operates under the same environmental conditions and with the same settings, the value of a parameter will lie within a band inside the accuracy window. Repeatability defines the width of this band.
Accuracy Window · Repeatability Band

Specifications Pulse Parameter Definitions





DATA ENTRY CURSOR/ DIGIT MODIFY
MAN
RECALL PERIOD DELAY WIDTH
STORE
AUTOSET
LEAD TRAIL HIGH LOW
HELP
ON | OFF
SHIFT
O . +/-
ENTER
LOCAL
DATA ENTRY / QUICK ACCESS
CURSOR
ENTER & UNIT Keys

Copyright Agilent Technologies 1998, 2000
Edition E0300 Printed in Germany
81101-91021


Manual for Model Number 81101A Manual printed on März 2000 E0300 Manual Part Number 81101-91021
Make all ERRATA corrections.
Check the following table for your instrument serial prefix/serial number/EDC and make the listed changes to your manual
Serial Prefix or Manual Serial Number Changes
DE389 00598 Serialnumber independent 1
5 September, 2001 Page 1 of 4
New Item

ERRATA
Page 91, Power consumption: 170VA max.
Page 92, Specifications change to read:
Recalibration period 3 years recommended
Page 93, Timing Specifications add:
The following specifications apply to all timing parameters unless otherwise specified in the following.
Repeatability: typically 4 times better than accuracy Resolution: 3.5 digits, best case 5 ps RMS Jitter: 0.01% + 15 ps
Page 93, Glitch-free timing changes change to read:
This applies to continous mode with timing values < 100 ms (frequency: > 10 Hz), and consecutive values between one-half and twice the previous value.
Page 94, Width add:
The pulse width is specified at fastest transitions.
Page 94, Delay add:
Delay and Double Pulse Delay are specified at fastest leading edge.
Page 97, Level Parameters add:
Page 99, Specifications add:
External Input:
Input Impedance: 50 Ohm or 1kOhm selectable Threshold: -10 V to + 10 V Maximum input voltage: +/- 15 V Input transitions: < 100 ns Input frequency: dc to max 50 MHz Minimum pulse width: 10 ns Input sensitivity:
05.09.01 Page 2 of 4
Common Specifications

INDEX OF MANUAL CHANGE
MANUAL CHANGE
FRAME
ERRATA
1 See attached Declaration of Conformity
05.09.01 Page 3 of 4

MANUAL CHANGE 1
On page 90, Specifications change to read: See attached Declarations of Conformity
05.09.01 Page 4 of 4

DECLARATION OF CONFORMITY According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014
Manufacturer’s Name: Agilent Technologies Deutschland GmbH Manufacturer’s Address: Boeblingen Verifications Solutions (BVS) Herrenberger Str. 130 D-71034 Boeblingen
Declares, that the product
Product Name: Family of Pulse-/Data Generators System Number: 81100 Product Modules: 81101A 50 MHz Pulse/Pattern Generator 81104A 80 MHz Pulse/Pattern Generator 81110A 330/165 MHz Pulse/Pattern Generator 81111A 165 MHz , 10 V Output Module 81112A 330 MHz , 3.5 V Output Module 81130A 400/660 MHz Pulse/Pattern Generator 81131A 400 MHz , 3.5 V Output Module 81132A 660 MHz , 2.5 V Output Module E8305A (a) VXI Plug-in 250 MHz Pulse Generator E8306A (a) VXI Plug-in 100 MHz Clock Generator E8311A (a) VXI Plug-in 165MHz Pulse/Pattern Generator E8312A (a) VXI Plug-in 330MHz Pulse/Pattern Generator
Conforms with the following European Directives:
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC and the EMC Directive 89/336/EEC (including 93/68/EEC) and carries the CE Marking accordingly.
Conforms with the following product standards:
EMC (Technical Construction File)
The product modules marked by (a) herewith comply with the requirements of the EMC Directive 89/336/EEC (including 93/68/EEC) and carry the CE Marking accordingly (European Union). Against: EMC test specification EN 55011:1991 (Group 1, Class A) As detailed in Electromagnetic Compatibility (EMC) Certificate of Compliance No. B801356L Assessed by: CETECOM ICT Services GmbH, D-66117 Saarbruecken
EMC IEC 61326-1:1997+A1:1998 / EN 61326-1:1997+A1:1998 CISPR 11:1997 / EN 55011:1998 Group 1 Class A (1) IEC 61000-4-2:1995+A1:1998 / EN 61000-4-2:1995 4kV CD, 8kV AD IEC 61000-4-3:1995 / EN 61000-4-3:1995 3 V/m, 80-1000 MHz IEC 61000-4-4:1995 / EN 61000-4-4:1995 0.5kV signal lines, 1kV power lines IEC 61000-4-5:1995 / EN 61000-4-5:1995 0.5 kV line-line, 1 kV line-ground IEC 61000-4-6:1996 / EN 61000-4-6:1996 3V, 0.15-80 MHz IEC 61000-4-11:1994 / EN 61000-4-11:1994 1 cycle/100% Canada: ICES-001:1998 Australia/New Zealand: AS/NZS 2064.1 Safety IEC 61010-1:1990+A1:1992+A2:1995 / EN 61010-1:1993+A2:1995 Canada: CSA C22.2 No. 1010.1:1992
Supplemental Information:
(1) The products were tested in a typical configuration with Agilent Technologies test systems.

Hans-Martin Fischer Date Name Product Regulations Engineer Title
For further information, please contact your local Agilent Technologies sales office, agent or distributor. Authorized EU-representative: Agilent Technologies Deutschland GmbH, Herrenberger Strasse 130, D-71034 Boeblingen, Germany
Standard Limit · 2001-May-02
Owners and our team usually reply within a day — and your question helps the next person too.
Be the first to ask — other owners or our team will answer.