Agilent Technologies Models 6811B - 6814B, 6834B, and 6843A AC Power Solutions


Agilent Part No. 5962-0883
Microfiche p/n 5962-0884 Printed in USA: April, 2000


14-character display shows menu commands and measured values. ♦ Annunciators indicate operating modes and status conditions.
Rotary controls set voltage and frequency when ac source is in local mode. ♦ Turn rapidly for coarse control ♦ Turn slowly for fine control.
ô Ì Ë
and scroll through menu commands.
and scroll through command parameters.
and scroll through lists and harmonic arrays.
ô index index

System keys: ♦ Return to Local mode ♦ Set the GPIB address and other system parameters ♦ Set the RS-232 interface ♦ Display SCPI error codes ♦ Save and recall instrument states
Function keys: ♦ Enable/disable the output ♦ Select output phases ♦ Select front panel metering and harmonic analysis functions ♦ Program voltage frequency, phase, current limit, pulse parameters, and waveform shapes ♦ Set and clear protection functions ♦ Select output and input coupling ♦ Monitor status ♦ Scroll through front panel menu commands.
Entry keys: ♦ Enter values ♦ Increment or decrement values ♦ Scroll through command parameters. ♦ Calibrate the ac source.


1 INH (Remote Inhibit) TTL input signal for externally disabling the power source. FLT (Discrete Fault Indicator) TTL output signal when there is a device fault. 2 RS-232 connector for remote controller. 3 TRIGGER BNC connectors for external trigger inputs and &source; trigger outputs. 4 GPIB connector and GPIB cable for remote controller. 5 SENSE connections for remote voltage sensing at the load. 6 Airflow Vents (do not block). 7 OUTPUT power connections to the load. (φ2, φ3 connections available on Agilent 6834B only.) 8 AC Input Line Fuses (Agilent 6814B/6834B/6843A only. Other models have internal fuses). 9 LINE RATING label specifies power source required by the power source. 10 AC Line Input connections from the power source.

Generate waveform shapes
Program the output
Agilent Models 6811B, 6812B, 6813B program the following additional output functions:
♦ Phase
♦ Ac rms voltage ♦ Dc voltage
♦ Distortion ♦ Peak Current limit
♦ Frequency ♦ AC coupling
♦ Voltage and frequency slew rates ♦ Impedance
♦ Rms current limit
Make the following measurements
Agilent Models 6811B, 6812B, 6813B make the following additional measurements:
♦ Ac rms, ac + dc rms voltage
♦ Ac rms, ac + dc rms current; plus repetitive and non-repetitive peak current
♦ Real, reactive, and apparent power ♦ Dc current
♦ Harmonic analysis of voltage and current waveforms giving amplitude, phase, and total harmonic distortion results of up to the 50th harmonic.
♦ Triggered acquisition of digitized voltage and current with extensive post-acquisition calculations
♦ Total power and neutral current


Operate under local or remote control
♦ From the front panel keys
♦ Through the built-in GPIB or RS-232 interfaces
Implement protection features
♦ Over-voltage
♦ Over-current
♦ Over-power
♦ Over-temperature
♦ User-defined external events (via a FLT shutdown signal)

Use these keys to move through the command menus of the selected function.

The following chart shows the commands in the Voltage function menu. Some commands may not appear on all models. Menus are circular, you can return to the starting position by continuously pressing p or q.
Use these keys to increment/decrement or select the command parameters to be executed. If the parameter is a number, use these keys to make minor changes to the value. Enter enters the selection and returns to the Meter function.
E
Key Display Description, , VOLT:M FIXED Sets fixed mode VOLT:M STEP Sets step mode VOLT:M PULSE Sets pulse mode VOLT:M LIST Sets list mode
to Use the numeric Entry keys to directly enter a value for the command parameter. For example, to enter a value for the voltage parameter:
Key Display Description VOLT 0 0 volts , VOLT 60 60 volts 60 V 60 Hz enters the value and returns to the Meter function

Make sure the unit is turned on. Use either the front panel keys or the corresponding SCPI commands.
The column on the left indicates the front panel keys that program the indicated action. If the SCPI programming syntax is substantially different from the front panel menu command, it is shown inside parentheses ( ).
The text to the right describes the result. If appropriate, the resultant output waveshape is shown underneath the description.
When the output is enabled, the programmed voltage appears at the output and the Dis annunciator turns off.
You can specify phases individually, or you can couple the phases. When phases are coupled, all three phase annunciators (φ1, φ2, φ3) on the front panel are on, indicating that commands will be sent to all three phases. Note that front panel metering is only done one phase at a time (except for the total power and neutral current measurements).
Set the voltage
When this command is sent, the output voltage is set to 120 V rms.
120 V rms 0 V
When this command is sent, the output frequency is set to 50 Hz.
60 Hz 50 Hz

When this command is sent, the rms current limit is set to 10 A. If more current than the programmed limit is drawn, the output voltage amplitude is reduced to keep the rms current within the specified limit. Press Shift Current and q to access CURR:PEAK, which lets you set the peak current limit on Agilent 6811B/6812B/6813B units. Note that the peak current limit circuit on these units acts instantly and clips the output voltage to maintain the programmed peak limit.
When this command is sent, the output generates a squarewave. Note that the peak-to-peak amplitude of the squarewave is less that that of a sinewave when it is programmed to the same rms voltage amplitude.
Program a protection function
These commands clear all previously set protection functions and then set the current protection, which disables the output when an overcurrent condition is detected. The OCP annunciator will light when this command is programmed.
CURR:PROT ON

All measurements are based on acquiring and subsequently processing output waveform information. When the ac source is on, it takes measurements and updates the front panel meter continuously. The Meter key accesses the measurement functions from the front panel.
The SCPI MEASure command acquires new waveform information each time it is executed. The SCPI FETCh command does not acquire new waveform information but extracts the desired information from previously acquired waveform data. SCPI commands let you measure phases individually or simultaneously measure all phases using the FETCh command.
The following example illustrates the measurements that can be returned by the front panel of the ac source when sourcing power to a typical non-resistive load such as a power supply. The ac source output voltage and current waveforms are shown on the next page.

1.93A 60HZ rms current and frequency
120V 150.5W rms voltage and power
2.82 CREST F current crest factor
5.379A PK REP peak current, repetitive
36.83A PK NR peak current, non repetitive

Use the harmonic menu to make harmonic measurements of the output current. The following example illustrates the current magnitude measurements returned at harmonics 0 to 5. Note that harmonic 1 is the fundamental. Harmonic 0 is the dc component.
1.43A I:MAG:1 current amplitude at harmonic 1
0.01A I:MAG:2 current amplitude at harmonic 2
0.91A I:MAG:3 current amplitude at harmonic 3
0.01A I:MAG:4 current amplitude at harmonic 4
0.74A I:MAG:5 current amplitude at harmonic 5
Output voltage and current waveforms

Up to now the ac source has been programmed with the transient system in Fixed mode. The following examples briefly describe the transient system’s Step, Pulse, and List modes, which require the application of a trigger to implement the transient mode.
Step transients transition to a new output level upon receipt of a trigger. When these commands are sent, the voltage amplitude is stepped from its previous setting to 150 V rms upon receipt of a trigger.
INIT IMMED
Program an output pulse
Pulse transients transition to a new output level upon receipt of a trigger and return to the original level after a specified time, repeating this action by the number of times specified by the count. When these commands are sent, two output pulses step the voltage amplitude from its previous setting to 90 V rms upon receipt of a trigger. At the end of the specified period (multiplied by the count), the voltage returns to its original level.
WIDTH .01 PER .03 COUNT 2
INIT IMMED
Program an output list
List transients generate complex output sequences. When these commands are sent, the voltage amplitude is sequentially stepped to three levels upon receipt of a trigger, and then returns to the original voltage level. The output remains at each list step for .5 seconds. The values inside the brackets ([ ]) are the list index references. Use Clear Entry to clear a list.
DWELL [0] .5 DWELL [1] .5 DWELL [2] .5 VOLT [0] 130 VOLT [1] 140 VOLT [2] 150 STEP AUTO

The previous examples showed how the transient system can be used to control the output voltage amplitude. The transient system can also control output frequency, phase, waveform shape, voltage and frequency slew rates, offset voltage, and peak current limit. The following examples illustrate how the transient system’s Pulse mode can generate frequency, shape, phase, and voltage slew pulses.

WIDTH .1
INIT IMMED
SHAPE:M PULSE SHAPE SINE SHAPE:T SQUARE
SHAPE TRIG
SHAPE
WIDTH .05
INIT IMMED
PHASE:M PULSE PHASE 0 PHASE:T 180
PHASE TRIG
PHASE
WIDTH .05
INIT IMMED
VOLT:M PULSE VOLT 120 VOLT:T 150 SLEW:M PULSE SLEW 10000 SLEW:T 1000

INIT IMMED

The previous transient examples were programmed to respond to immediate triggers. However, delayed and phase synchronized triggers can also be programmed as shown in the following examples.
When these commands are sent, the voltage amplitude changes immediately upon the receipt of a trigger.
DELAY 0 SYNC:SOUR IMM INIT:IMMED
No delay; 90 degrees phase synchronization
When these commands are sent, the voltage amplitude changes at the next 90 degree phase angle that occurs following the receipt of a trigger.
DELAY 0 SYNC:SOUR PHAS SYNC:PHAS 90 INIT:IMMED
Trigger delay; no phase synchronization
When these commands are sent, the voltage amplitude changes .0167 seconds after the receipt of a trigger.
DELAY .0167 SYNC:SOUR IMM INIT:IMMED

When these commands are sent, the voltage amplitude changes at the next 90 degree phase angle that occurs after the .0167 second delay has expired, following the receipt of a trigger.
DELAY .0167 SYNC:SOUR PHAS SYNC:PHAS 90 INIT:IMMED
More about the trigger system
In the previous examples, a front panel trigger is used to generate the output transients. The trigger is shown occurring at 270 degrees but actual triggers may occur at any phase. Delay and phase synchronization however, will occur as programmed.
INIT:IMMED
INITIATED STATE
TRIGGER RECEIVED
DELAYING STATE
DELAY COMPLETED
WAIT FOR SYNC STATE
PHASE ANGLE OCCURS
OUTPUT CHANGES


Press to change the ac source’s selected interface from remote operation to local (front panel) operation. Pressing the key will have no effect if the interface state is already Local, Local- with-Lockout, or Remote-with-Lockout. Error
Meter Functions V Hz rms voltage and frequency V A rms voltage and rms current A Hz rms current and frequency V W rms voltage and power CREST F current crest factor A PK REP peak current, repetitive A PK NR peak current, nonrepetitive VA apparent power VAR reactive power W TOTAL total power all phases1 PFACTOR power factor A NEUTRAL neutral rms current1

Error Functions ERROR Displays system error codes stored in the SCPI error queue. If no errors exist, a 0 is displayed. The Err annunciator is lit when there are errors.
Address Functions ADDRESS sets the GPIB address INTF GPIB | RS232 selects an interface BAUDRATE 300 | 600 | 1200
PARITY NONE | EVEN | ODD selects message parity LANG SCPI | E9012 selects the language NOUTPUTS 1 | 3 selects the number of outputs1
Output Functions OUTP:COUP AC | DC select output coupling3
*RST executes the *RST command TTLT:SOUR BOT| EOT| LIST select Trigger Out source coupling TTLT:STATE ON | OFF set Trigger Out state IMP:STATE ON | OFF set output impedance programming3 IMP:REAL set real part of output impedance3 IMP:REAC set reactive part of output impedance3 PON:STATE RST | RCL0 select power-on state command RI LATCHING | LIVE | OFF sets remote inhibit mode DFI ON | OFF sets discrete fault indicator state DFI:SOUR QUES | OPER

Save Functions Press to save an existing ac source state in nonvolatile memory. Up to 16 states can be saved (0-15).
Recall Functions Press to place the ac source into a previously saved state. Up to 16 states can be recalled (0-15).
select the DFI source
First press and release this blue shift key to select a shifted function. The Shift annunciator lights when this key is pressed.
Input Functions INP:COUP AC | DC | ACDC choose meter coupling CURR:RANGE HIGH | LOW current measurement range3 WINDOW KBESSEL | RECT select harmonic measurement
Status Functions *CLS executes the *CLS command STATUS:PRESET executes STATus:PRESet command *ESR? return Event Status register value *STB return Status Byte register value OPER:EVEN? return STAT:OPER:EVENT? value OPER:COND return STAT:OPER:COND? value QUES:EVEN? return STAT:QUES:EVENT? value QUES:COND return STAT:QUES:COND? value Protect Functions PROT:CLEAR clears latched protection signal CURR:PROT ON | OFF set overcurrent protection function VOLT:PROT ON | OFF set overvoltage protection function3 VOLT:PROT set overvoltage protection level DELAY set a time delay for activating a protection fault

Harmonic Functions A I:MAG: current harmonic magnitude ° I:PHASE: current harmonic phase V V:MAG: voltage harmonic magnitude ° V:PHASE: voltage harmonic phase N:MAG: neutral current harmonic magnitude ° N:PHASE: neutral current harmonic phase CURR:THD current total % harmonic distortion VOLT:THD voltage total % harmonic distortion
Meter functions continued on next column



Trigger Function Pressing the Shift Trigger key generates an immediate trigger
Trigger Control Functions INIT:IMMED Initiate trigger immediately INIT:CONT ON | OFF Initiate trigger continuously TRIG:SOUR BUS | EXT
DELAY Set trigger delay in seconds ABORT Abort all trigger sequences SYNC:SOUR PHASE | IMM Select synchronous trigger source SYNC:PHASE Set synchronous phase reference

List Functions COUNT number of times a list repeats DWEL: list of output dwell times FREQ: list of output frequencies FSLW: list of output frequency slew rates IPK: list of output peak current limits3 OFFS: list of dc output voltages3 OSLW: dc offset voltage slew rate list3 PHASE: list of output voltage phase angles4 SHAP: SINE | SQUARE
Current Functions CURR:LEV set immediate rms current limit4
Voltage Functions VOLT set immediate ac output voltage4
list of output waveform shapes
STEP ONCE | AUTO set response of list to triggers TTLT: ON | OFF set trigger out pulse list VOLT: list of ac output voltages4 VSLW: list of output voltage slew rates4 Pulse Functions WIDTH set the pulse width COUNT set the number of output pulses DCYCLE set the pulse duty cycle PER set the pulse period count HOLD WIDTH | DCYCLE set parameter that is held constant
RANGE 150 | 300 set the voltage range2, 4
OFF:SLW set dc offset slew in V/sec3
OFF:SLW:T set triggered dc offset slew in V/sec3
q Index p Index

ALC INT | EXT select the voltage sense source ALC:DET RTIME | RMS select the voltage sense detector3
q p Index Functions These are Shift Index keys which are used to scroll through indexed functions. Press these keys to step through integers 0 through 50 for a harmonic list, or 0 through 99 for list points. Hold down these keys to rapidly access any harmonic or list point.

q p Functions These keys let you move through the choices in a command list. Command lists are circular; you can return to the starting position by continuously pressing either key.
Phase Functions PHASE set immediate output phase4
Freq Functions FREQ set immediate output frequency FREQ:T set triggered output frequency FREQ:M FIXED | STEP
This key applies to 3-phase ac sources only. Pressing this key successively selects phase 1 first, followed by phase 2, phase 3, and then all three phases.
This key toggles the output on and off. When off, the ac source output is disabled and the Dis annunciator is on.
SLEW set frequency slew in Hz/sec SLEW:T set triggered frequency slew Hz/sec SLEW:M FIXED | STEP

These keys let you scroll through choices in a parameter list that apply to a specific command. Parameter lists are circular; you can return to the starting position by continuously pressing either key. If the command has a numeric range, these keys increment or decrement the existing value.
− The numeric keys 0 through 9 are used for entering numeric values. − Press shift and this key to enter a minus. Press this key alone to enter a decimal point.
Until you press the Enter key, the values or parameters you enter with the other Entry keys are displayed but not entered into the ac source. E Press Shift and this key to enter an exponent.
Clear Entry Press Shift and this key to abort a keypad entry and clear the value. When editing a list, pressing Clear Entry truncates or clears the list at the presently displayed list point. Press this key alone to backspace and delete the last digit entered
Calibration Press Shift and this key to access the calibration menu. Refer to appendix B In the user’s for more information.
¯ ° · Enter

For more information about Agilent Technologies test and measurement products, applications, services, and for a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir
You can also contact one of the following centers and ask for a test and measurement sales representative.
United States: Agilent Technologies Test and Measurement Call Center P.O. Box 4026 Englewood, CO 80155-4026 (tel) 1 800 452 4844
Latin America: Agilent Technologies Latin American Region Headquarters 5200 Blue Lagoon Drive, Suite #950 Miami, Florida 33126 U.S.A. (tel) (305) 267 4245 (fax) (305) 267 4286
Canada: Agilent Technologies Canada Inc. 5150 Spectrum Way Mississauga, Ontario L4W 5G1 (tel) 1 877 894 4414
Australia/New Zealand: Agilent Technologies Australia Pty Ltd 347 Burwood Highway Forest Hill, Victoria 3131 (tel) 1-800 629 485 (Australia) (fax) (61 3) 9272 0749 (tel) 0 800 738 378 (New Zealand) (fax) (64 4) 802 6881
Europe: Agilent Technologies Test & Measurement European Marketing Organisation P.O. Box 999 1180 AZ Amstelveen The Netherlands (tel) (31 20) 547 9999
Asia Pacific: Agilent Technologies 24/F, Cityplaza One, 1111 King’s Road, Taikoo Shing, Hong Kong tel: (852)-3197-7777 fax: (852)-2506-9284
Japan: Agilent Technologies Japan Ltd. Measurement Assistance Center 9-1, Takakura-Cho, Hachioji-Shi, Tokyo 192-8510, Japan (tel) (81) 426 56 7832 (fax) (81) 426 56 7840
Technical data is subject to change.