The Agilent 1680/90-Series Logic Analyzer-At a Glancep.2Featuresp.2Service Strategyp.3In This Bookp.4General Informationp.9Accessoriesp.10Accessories Availablep.10Specificationsp.11Characteristicsp.11Probesp.12Auxiliary Powerp.12Operating Environment (for indoor use only)p.12Dimensionsp.13Recommended Test Equipmentp.14Preparing for Usep.15Power Requirementsp.16Operating Environmentp.16Storagep.16To inspect the logic analyzerp.16To apply powerp.17To connect the 1690A,AD-series logic analyzer to a host PCp.17To start the user interfacep.18To clean the logic analyzerp.18To test the logic analyzerp.18Testing Performancep.19The Logic Analyzer Interfacep.20Test Strategyp.20Test Intervalp.20Performance Test Recordp.20Test Equipmentp.20To make the test connectorsp.21To set up the test equipment and the logic analyzerp.23Set up the test equipmentp.23Set up the 1680A,AD-series logic analyzerp.24Set up the 1690A,AD-series logic analyzerp.25To perform the logic analyzer self-testsp.25To test the threshold accuracyp.28Set up the equipmentp.28Connect and configure the logic analyzerp.29Test the ECL Thresholdp.30Test the 0 V User Thresholdp.31Test the next podp.32To set up the logic analyzer for the state mode testsp.33To test the single-clock, single-edge, state acquisitionp.37Set up the equipmentp.37Connect and configure the logic analyzerp.37Verify the test signalp.40Check the setup/hold combinationp.41Test the next channels (1680/81A,AD and 1690/91A,AD)p.47To test the multiple-clock state acquisitionp.48Set up the equipmentp.48Connect and configure the logic analyzerp.49Verify the test signalp.51Check the setup/hold with single clock edges, multiple clocksp.52Test the next channels (1680/81A, AD and 1690/91A, AD)p.56To test the single-clock, multiple-edge, state acquisitionp.57Set up the equipmentp.57Connect and configure the logic analyzerp.58Verify the test signalp.60Check the setup/hold with single clock, multiple clock edgesp.61Test the next channels (1680/81A,AD and 1690/91A,AD)p.65To test the time interval accuracyp.66Set up the equipmentp.66Connect and configure the logic analyzerp.67Acquire and verify the test datap.69Performance Test Recordp.71Calibrating and Adjustingp.75Logic analyzer calibrationp.76Troubleshootingp.77To install the fan guardp.78To use the flowchartsp.79Troubleshooting the Agilent 1680A,AD-seriesp.80To check the power-up testsp.87To test the power supply voltagesp.87To test the LCD display signalsp.89To test disk drive voltagesp.90To verify the CD-ROMp.92To recover the operating systemp.93To reinstall the operating systemp.93Problems running the Application Softwarep.93Problems with the Operating Systemp.94Troubleshooting the Agilent 1690A,AD-seriesp.95To verify connectivityp.99Task Managerp.99Device Managerp.100To test the power supply voltagesp.100General Troubleshootingp.103To run the self-testsp.103Acquisition board status LEDsp.106Green LEDsp.106Red LEDp.106Normal Operationp.106To test the logic analyzer probe cablesp.107To check the BNC Trigger input/output signalsp.110To test the auxiliary powerp.112Replacing Assembliesp.1131680A,AD-series disassembly/assemblyp.114Prepare the instrument for disassemblyp.114To remove the chassis from the sleevep.114To remove the acquisition boardp.115To remove the power supplyp.117To remove the hard disk drivep.118To remove the CD-ROM drive assemblyp.119Remove the CD-ROM from the assemblyp.119To remove the flexible disk drivep.120To remove the PCI boardsp.122To remove the motherboardp.123Transfer the I/O panel to the replacement boardp.124To remove the front panel assemblyp.126To disassemble the front panel assemblyp.128Remove the front panel circuit board.p.128Remove the LCD display.p.128Remove the on/off switch.p.128To remove the distribution boardp.129To remove the inverter boardp.130To remove the fansp.131To remove the cable trayp.1321690A,AD-series disassembly/assemblyp.133Prepare the instrument for disassemblyp.133To remove the chassis from the sleevep.133To remove the fasciap.134Steps to remove the on/off switch from the fascia assemblyp.135To remove the acquisition boardp.136Steps to remove the probe shroud from the acquisition board.p.136To remove the deckp.137To remove the power supplyp.137To remove the distribution boardp.138To remove the fansp.139To remove the line filterp.140To remove the front panel and front framep.141Replaceable Partsp.143Replaceable Parts Orderingp.144Parts listedp.144Parts not listedp.144Direct mail order systemp.144Exchange Assembliesp.145Replaceable Parts Listp.145Exploded Viewp.146Agilent 1680A,AD-Series Replaceable Partsp.147Exploded Viewp.153Agilent 1690A,AD-Series Replaceable Partsp.154Power Cables and Plug Configurationsp.157Theory of Operationp.159Block-Level Theoryp.160Agilent 1680A,AD-series Logic Analyzer Theoryp.161PC Motherboardp.161PC Software Systemp.161Disk Drivesp.161Flat Panel Displayp.161Networkingp.161Other I/Op.161Power Supplyp.162Acquisition Boardp.162Probingp.162Comparatorsp.163Acquisitionp.163Acquisition RAMp.163Test and Clock Synchronization Circuitp.164Clock and Data Thresholdp.164Motherboard Interfacep.164Memory Address Counters (MACs)p.164+5 VDC supplyp.164Power Distribution Boardp.165Front Panel Boardp.165Agilent 1690A,AD-series Logic Analyzer Theoryp.166Acquisition Boardp.166Power Supplyp.166Power Distribution Boardp.166Self-Tests Descriptionsp.167Power-up Self-Tests (1680A,AD-series)p.167Connectivity Tests (1690A,AD-series)p.167Acquisition Board Self Testsp.168Register Testp.168Memory Testp.168Comparator Testp.168Trigger Bus Testp.168Trigger Arm Testp.169Clock Paths Testp.169Memory Modes Testp.169Calibration Testp.169Logic Analyzer Self-Testsp.169Register Testp.169Memory Testp.170Comparator Testp.170Trigger Bus Testp.170Trigger Arm Testp.170Clock Paths Testp.170Memory Modes Testp.171Calibration Testp.171
Page 1 of the Agilent Technologies 1690 user manual — Agilent Technologies 1680/90-Series Logic Analyzer
Page 1 · click to enlarge
Page 1

Service Guide

Publication number 01680-97016 October 2005

For Safety information, Warranties, and Regulatory information, see the pages at the end of the book.

© Copyright Agilent Technologies 2001, 2004-2005

All Rights Reserved.

Agilent Technologies 1680/90-Series Logic Analyzer

Page 2 of the Agilent Technologies 1690 user manual — The Agilent 1680/90-Series Logic Analyzer–At a Glance
Page 2 · click to enlarge
Page 2

The Agilent 1680/90-Series Logic Analyzer–At a Glance

Features

Some of the main features of the Agilent 1680A,AD-Series Logic Analyzers are as follows:

  • Standalone benchtop logic analyzer
  • Microsoft Windows® XP Professional operating system
  • 132 data channels and 4 clock/data channels on the Agilent 1680A,AD
  • 98 data channels and 4 clock/data channels on the Agilent 1681A,AD
  • 64 data channels and 4 clock/data channels on the Agilent 1682A,AD
  • 32 data channels and 2 clock/data channels on the Agilent 1683A,AD
  • 12.1-inch LCD display
  • 3.5-inch flexible disk drive
  • 80GB hard disk drive
  • Centronics and LAN interfaces
  • IEEE 1394 interface for hosted control
  • Variable setup/hold time
  • 512K acquisition memory in the Agilent 1680A-series
  • 2M acquisition memory in the Agilent 1680AD-series
  • Marker Measurements
  • PS/2 Mouse
  • PS/2 keyboard support

Some of the main features of the Agilent 1690A,AD-Series Logic Analyzers are as follows:

  • Hosted benchtop logic analyzer
  • 132 data channels and 4 clock/data channels on the Agilent 1690A,AD
  • 98 data channels and 4 clock/data channels on the Agilent 1691A,AD
  • 64 data channels and 4 clock/data channels on the Agilent 1692A,AD
  • 32 data channels and 2 clock/data channels on the Agilent 1693A,AD
  • IEEE 1394 interface for hosted control
  • Variable setup/hold time
  • 512K acquisition memory in the Agilent 1690A-series
  • 2M acquisition memory in the Agilent 1690AD-series
  • Marker Measurements
Page 3 of the Agilent Technologies 1690 user manual — Service Strategy
Page 3 · click to enlarge
Page 3

Service Strategy

The service strategy for this instrument is the replacement of defective assemblies. This service guide contains information for finding a defective assembly by testing and servicing the Agilent 1680/90-series logic analyzers.

This logic analyzer can be returned to Agilent for all service work, including troubleshooting. Contact your nearest Agilent Technologies Sales Office for details.

Agilent Technologies 1690 — Service Strategy diagram (page 3)
Fig. 1 · Service Strategy

Agilent Technologies 1680-Series Logic Analyzer

Agilent Technologies 1690 — Service Strategy diagram (page 3)
Fig. 2 · Service Strategy

Agilent Technologies 1690-Series Logic Analyzer

Page 4 of the Agilent Technologies 1690 user manual — In This Book
Page 4 · click to enlarge
Page 4

In This Book

This book is the service guide for the Agilent 1680/90-Series Logic Analyzers and is divided into eight chapters.

Chapter 1 contains information about the logic analyzer and includes accessories, specifications and characteristics, and equipment required for servicing.

Chapter 2 tells how to prepare the logic analyzer for use.

Chapter 3 gives instructions on how to test the performance of the logic analyzer.

Chapter 4 contains calibration instructions for the logic analyzer.

Chapter 5 contains self-tests and flowcharts for troubleshooting the logic analyzer.

Chapter 6 tells how to replace assemblies of the logic analyzer and how to return them to Agilent Technologies.

Chapter 7 lists replaceable parts, shows an exploded view, and gives ordering information.

Chapter 8 explains how the logic analyzer works and what the self-tests are checking.

Page 5 of the Agilent Technologies 1690 user manual — 3 Testing Performance
Page 5 · click to enlarge
Page 5

Contents

The Agilent 1680/90-Series Logic Analyzer–At a Glance

Features 2 Service Strategy 3

In This Book

1 General Information

Accessories 10 Specifications 11 Characteristics 11 Recommended Test Equipment 14

2 Preparing for Use

Power Requirements 16 Operating Environment 16 Storage 16 To inspect the logic analyzer 16 To apply power 17 To connect the 1690A,AD-series logic analyzer to a host PC 17 To start the user interface 18 To clean the logic analyzer 18 To test the logic analyzer 18

3 Testing Performance

The Logic Analyzer Interface 20 Test Strategy 20 Test Interval 20 Performance Test Record 20 Test Equipment 20

To make the test connectors 21

To set up the test equipment and the logic analyzer 23 Set up the test equipment 23 Set up the 1680A,AD-series logic analyzer 24 Set up the 1690A,AD-series logic analyzer 25 To perform the logic analyzer self-tests 25

Page 6 of the Agilent Technologies 1690 user manual — 4 Calibrating and Adjusting
Page 6 · click to enlarge
Page 6

Contents

To test the threshold accuracy 28 Set up the equipment 28 Connect and configure the logic analyzer 29 Test the ECL Threshold 30 Test the 0 V User Threshold 31 Test the next pod 32

To set up the logic analyzer for the state mode tests 33

To test the single-clock, single-edge, state acquisition 37 Set up the equipment 37 Connect and configure the logic analyzer 37 Verify the test signal 40 Check the setup/hold combination 41 Test the next channels (1680/81A,AD and 1690/91A,AD) 47

To test the multiple-clock state acquisition 48 Set up the equipment 48 Connect and configure the logic analyzer 49 Verify the test signal 51 Check the setup/hold with single clock edges, multiple clocks 52 Test the next channels (1680/81A, AD and 1690/91A, AD) 56

To test the single-clock, multiple-edge, state acquisition 57 Set up the equipment 57 Connect and configure the logic analyzer 58 Verify the test signal 60 Check the setup/hold with single clock, multiple clock edges 61 Test the next channels (1680/81A,AD and 1690/91A,AD) 65

To test the time interval accuracy 66 Set up the equipment 66 Connect and configure the logic analyzer 67 Acquire and verify the test data 69

Performance Test Record 71

4 Calibrating and Adjusting

Logic analyzer calibration 76

Page 7 of the Agilent Technologies 1690 user manual — 6 Replacing Assemblies
Page 7 · click to enlarge
Page 7

Contents

5 Troubleshooting

To install the fan guard 78 To use the flowcharts 79

Troubleshooting the Agilent 1680A,AD-series 80 To check the power-up tests 87 To test the power supply voltages 87 To test the LCD display signals 89 To test disk drive voltages 90 To verify the CD-ROM 92 To recover the operating system 93

Troubleshooting the Agilent 1690A,AD-series 95 To verify connectivity 99 To test the power supply voltages 100

General Troubleshooting 103 To run the self-tests 103 Acquisition board status LEDs 106 To test the logic analyzer probe cables 107 To check the BNC Trigger input/output signals 110 To test the auxiliary power 112

6 Replacing Assemblies

1680A,AD-series disassembly/assembly 114 Prepare the instrument for disassembly 114 To remove the chassis from the sleeve 114 To remove the acquisition board 115 To remove the power supply 117 To remove the hard disk drive 118 To remove the CD-ROM drive assembly 119 To remove the flexible disk drive 120 To remove the PCI boards 122 To remove the motherboard 123 To remove the front panel assembly 126 To disassemble the front panel assembly 128 To remove the distribution board 129 To remove the inverter board 130 To remove the fans 131 To remove the cable tray 132

Page 8 of the Agilent Technologies 1690 user manual — 8 Theory of Operation
Page 8 · click to enlarge
Page 8

Contents

1690A,AD-series disassembly/assembly 133 Prepare the instrument for disassembly 133 To remove the chassis from the sleeve 133 To remove the fascia 134 To remove the acquisition board 136 To remove the deck 137 To remove the power supply 137 To remove the distribution board 138 To remove the fans 139 To remove the line filter 140 To remove the front panel and front frame 141

7 Replaceable Parts

Replaceable Parts Ordering 144 Replaceable Parts List 145 Exploded View 146 Agilent 1680A,AD-Series Replaceable Parts 147 Exploded View 153 Agilent 1690A,AD-Series Replaceable Parts 154 Power Cables and Plug Configurations 157

8 Theory of Operation

Agilent 1680A,AD-series Logic Analyzer Theory 161 Power Supply 162 Acquisition Board 162 Power Distribution Board 165 Front Panel Board 165

Agilent 1690A,AD-series Logic Analyzer Theory 166 Acquisition Board 166 Power Supply 166 Power Distribution Board 166

Self-Tests Descriptions 167 Power-up Self-Tests (1680A,AD-series) 167 Connectivity Tests (1690A,AD-series) 167 Acquisition Board Self Tests 168 Logic Analyzer Self-Tests 169

Block-Level Theory 160

Page 9 of the Agilent Technologies 1690 user manual — General Information
Page 9 · click to enlarge
Page 9

General Information

This chapter lists the accessories, the specifications and characteristics, and the recommended test equipment.

Page 10 of the Agilent Technologies 1690 user manual — Accessories Available
Page 10 · click to enlarge
Page 10
Chapter 1General Information

Accessories

The following accessory is supplied with the Agilent 1680/90-series logic analyzers. The part number is current as of the print date of this edition of the Service Guide, but further upgrades may change the part number. Do not be concerned if the accessory you receive has a different part number.

Accessories Supplied Agilent Part Number Qty

3-button Corded Mouse 1150-7845 1 Mini Keyboard 1150-7809 1

Accessories Available

For probing information, see Probing Solutions for Logic Analysis Systems (publication 5968-4632E) available at:

http://www.agilent.com/find/logic_analyzer_probes

Page 11 of the Agilent Technologies 1690 user manual — Characteristics
Page 11 · click to enlarge
Page 11
Chapter 1General Information

Specifications

The specifications are the performance standards against which the product is tested.

Maximum State Speed (selectable) 200 MHz Minimum Master to Master Clock Time* 5.000 ns Threshold Accuracy ± (65 mV + 1.5% of threshold setting)

Single Clock, Single Edge 4.5/-2.0 ns through -2.0/4.5 ns, adjustable in 100 ps increments Single Clock, Multiple Edges 5.0/-2.0 ns through -1.5/4.5 ns, adjustable in 100 ps increments Multiple Clocks, Multiple Edges 5.0/-2.0 ns through -1.5/4.5 ns, adjustable in 100 ps increments * Specified for an input signal VH = −0.9 V, VL = −1.7 V, slew rate = 1 V/ns, and threshold = −1.3 V.

Characteristics

These characteristics are not specifications, but are included as additional information.

Full Channel Half Channel

Maximum State Clock Rate 150 MHz not applicable Maximum Conventional Timing Rate 250 MHz 500 MHz Memory Depth (1680A, or 1690A-series) 512 K 1024 K Memory Depth (1680AD or 1690AD-series) 2048 K 4196 K

1680A,AD or 1690A,AD 136 68 1681A,AD or 1691A,AD 102 51 1682A,AD or 1692A,AD 68 34 1683A,AD or 1693A,AD 34 17

Setup/Hold Time* · Channel Count

Page 12 of the Agilent Technologies 1690 user manual — Operating Environment (for indoor use only)
Page 12 · click to enlarge
Page 12
Chapter 1General Information

Probes

Maximum Input Voltage + 40V Peak AC+DC, CAT 1

Auxiliary Power

Power Through Cables 1/3 amp at 5 V maximum per cable, CAT 1

Operating Environment (for indoor use only)

Temperature:

  • Instrument: 0 ° C to 55 ° C (+32 ° F to 131 ° F).
  • Probe lead sets and cables: 0 ° C to 65 ° C (+32 ° F to 149 ° F).
  • Disk media: 10 ° C to 40 ° C (+50 ° F to 104 ° F).

Humidity: Instrument, probe lead sets, and cables, up to 80% relative humidity at +40° C (+122° F).

Altitude: To 3067 m (10,000 ft).

Vibration:

  • Operating: Random vibration 5 to 500 Hz, 10 minutes per axis, 0.3 g (rms).
  • Non-operating: Random vibration 5 to 500 Hz, 10 minutes per axis, 2.41 g (rms); and swept sine resonant search, 5 to 500 Hz, 0.75 g (0-peak), 5 minute resonant dwell at 4 resonances per axis.
Page 13 of the Agilent Technologies 1690 user manual — Dimensions
Page 13 · click to enlarge
Page 13
Chapter 1General Information

Dimensions

Agilent Technologies 1690 — Dimensions diagram (page 13)
Fig. 3 · Dimensions
Agilent Technologies 1690 — Dimensions diagram (page 13)
Fig. 4 · Dimensions
Agilent Technologies 1690 — Dimensions diagram (page 13)
Fig. 5 · Dimensions

1680A,AD-Series · 1690A,AD-Series

Page 14 of the Agilent Technologies 1690 user manual — 10503A T
Page 14 · click to enlarge
Page 14
Chapter 1General Information

Recommended Test Equipment

Equipment Critical Specifications Recommended Model/Part Use*

Pulse Generator 200 MHz, 2.5 ns pulse width,

8133A Option 003 P,T

Digitizing Oscilloscope ≥6 GHz bandwidth,

54750A mainframe with 54751A plug-in module

P

Function Generator Accuracy (5)(10-6) frequency, DC offset voltage ±1.6 V

33250A P

Digital Multimeter 0.1 mV resolution, 0.005% accuracy

3458A P

BNC-Banana Cable 11001-60001 P

BNC Tee BNC (m)(f)(f) 1250-0781 P

Cable BNC (m)(m) 48 inch 8120-1840 P,T

SMA Coax Cable (Qty 3) 18 GHz bandwidth 8120-4948 P

Adapter (Qty 4) SMA(m)-BNC(f) 1250-1200 P, T

Adapter SMA(f)-BNC(m) 1250-2015 P

Coupler (Qty 4) BNC (m)(m) 1250-0216 P, T

20:1 Probes (Qty 2) 54006A P

BNC Coax Cable BNC(m)(m), >2 GHz bandwidth, >1 meter length

10503A T

BNC Test Connector, 17x2 (Qty 1)** P

BNC Test Connector, 6x2 (Qty 4)** P,T

Digitizing Oscilloscope > 100 MHz Bandwidth 54600B T

BNC Shorting Cap 1250-0774 T

BNC-Banana Adapter 1251-2277 T

*A = Adjustment P = Performance Tests T = Troubleshooting **Instructions for making these test connectors are in chapter 3, "Testing Performance."

Equipment Required

Page 15 of the Agilent Technologies 1690 user manual — Preparing for Use
Page 15 · click to enlarge
Page 15

Preparing for Use

This chapter gives you instructions for preparing the logic analyzer for use.

Page 16 of the Agilent Technologies 1690 user manual — To inspect the logic analyzer
Page 16 · click to enlarge
Page 16
Chapter 2Preparing for Use

Power Requirements

The logic analyzer requires a power source of either 115 Vac or 230 Vac, –22 % to +10 %, single phase, 48 to 66 Hz, CAT II pollution degree 2, 140/400 Watts nominal maximum power (1680A/AD-series), and 76/200 Watts nominal maximum power (1690A/AD-series).

Operating Environment

The operating environment is listed in chapter 1. The logic analyzer will operate at all specifications within the temperature and humidity range given in chapter 1. However, reliability is enhanced when operating the logic analyzer within the following ranges:

  • Temperature: +20 ° C to +35 ° C (+68 ° F to +95 ° F)
  • Humidity: 20% to 80% noncondensing

Storage

Store or ship the logic analyzer in environments within the following limits:

  • Temperature: -40 ° C to +75 ° C
  • Humidity: Up to 90% at 65 ° C
  • Altitude: Up to 15,300 meters (50,000 feet)

Protect the logic analyzer from temperature extremes which cause condensation on the instrument.

To inspect the logic analyzer

1 Inspect the shipping container for damage.

If the shipping container or cushioning material is damaged, keep them until you have checked the contents of the shipment and checked the instrument mechanically and electrically.

2 Check the supplied accessories.

Accessories supplied with the logic analyzer are listed in “Accessories” on page 10.

Page 17 of the Agilent Technologies 1690 user manual — 128MB RAM
Page 17 · click to enlarge
Page 17
Chapter 2Preparing for Use

3 Inspect the product for physical damage.

Check the logic analyzer and the supplied accessories for obvious physical or mechanical defects. If you find any defects, contact your nearest Agilent Technologies Sales Office. Arrangements for repair or replacement are made, at Agilent Technologies' option, without waiting for a claim settlement.

To apply power

These steps are required for all 1680A,AD and 1690A,AD-series logic analyzers.

1 Connect the power cord to the instrument and to the power source.

This instrument autodetects the line voltage from 115 VAC to 230 VAC. It is equipped with a three-wire power cable. When connected to an appropriate AC power outlet, this cable grounds the instrument cabinet. The type of power cable plug shipped with the instrument depends on the country of destination. Refer to chapter 7, "Replaceable Parts," for option numbers of available power cables.

2 Turn on the power switch located on the front panel.

To connect the 1690A,AD-series logic analyzer to a host PC

These steps are required for the Agilent 1690A,AD-series hosted logic analyzer.

The logic analyzer user interface requires a host computer (PC) with the following characteristics (or better):

Intel Celeron, AMD K6-II 500 MHz

Windows 2000 Professional or Windows XP Professional

128MB RAM

IEEE 1394 PCI card

1 Connect one end of the 6-pin IEEE 1394 cable to the IEEE 1394 port on the host PC.

2 Connect the free end of the IEEE 1394 cable to the IEEE 1394 port on the logic analyzer.

3 Apply power to the PC if it is not turned on.

Page 18 of the Agilent Technologies 1690 user manual — To test the logic analyzer
Page 18 · click to enlarge
Page 18
Chapter 2Preparing for Use

To start the user interface

Start the Agilent Logic Analyzer application from the Start menu or using a shortcut. On the desktop, the Agilent Logic Analyzer icon looks like:

Refer to the Agilent Logic Analyzer on-line help for information on how to operate the user interface. Also, refer to the window icon reference on the inside front cover of this service manual for a brief explanation of the Agilent Logic Analyzer standard icons.

To clean the logic analyzer

With the instrument turned off and unplugged, use mild soap and water to clean the front and cabinet of the logic analyzer. Harsh soap might damage the water- base paint.

To test the logic analyzer

  • If you require a test to verify the specifications, start at the beginning of chapter 3, "Testing Performance."
  • If you require a test to initially accept the operation, perform the self-tests in chapter 3.
  • If the logic analyzer does not operate correctly, go to the beginning of chapter 5, "Troubleshooting."
Page 19 of the Agilent Technologies 1690 user manual — Testing Performance
Page 19 · click to enlarge
Page 19

Testing Performance

This chapter tells you how to test the performance of the logic analyzer against the specifications listed in chapter 1.

Page 20 of the Agilent Technologies 1690 user manual — Test Equipment
Page 20 · click to enlarge
Page 20
Chapter 3Testing Performance

The Logic Analyzer Interface

To select a field on the logic analyzer screen, use the arrow keys to highlight the field, then press the Select key. Provided on the inside front cover of this manual is a list of logic analyzer icons that can be referenced while performing test procedures. For more information about the logic analyzer interface, refer to the Agilent Logic Analyzer application’s online help.

Test Strategy

For a complete test, start at the beginning with the software tests and continue through to the end of the chapter. For an individual test, follow the procedure in the test. The examples in this chapter were performed using an Agilent 1680AD. Other analyzers in the series will have appropriate pods showing on the screen.

The performance verification procedures starting on page 3–8 are each shown from power-up. To exactly duplicate the setups in the tests, save the power-up configuration to a file on a disk, then load that file at the start of each test.

If a test fails, check the test equipment setup, check the connections, and verify adequate grounding. If a test still fails, the most probable cause of failure would be the acquisition board.

Test Interval

Test the performance of the logic analyzer against specifications at two-year intervals.

Performance Test Record

A performance test record for recording the results of each procedure is located at the end of this chapter. Use the performance test record to gauge the performance of the logic analyzer over time.

Test Equipment

Each procedure lists the recommended test equipment. You can use equipment that satisfies the specifications given. However, the procedures are based on using the recommended model or part number. Before testing the performance of the logic analyzer, warm-up the instrument and the test equipment for 30 minutes.

Page 21 of the Agilent Technologies 1690 user manual — To make the test connectors
Page 21 · click to enlarge
Page 21
Chapter 3Testing Performance

To make the test connectors

To make the test connectors

The test connectors connect the logic analyzer to the test equipment.

Description Recommended Part Qty

BNC (f) Connector 1250-0698 5

100 Ω 1% resistor 0698-7212 8

Berg Strip, 17-by-2 1

Berg Strip, 6-by-2 4

20:1 Probe 54006A 2

1 Build four test connectors using BNC connectors and 6-by-2 sections of Berg strip:

a Solder a jumper wire to all pins on one side of the Berg strip.

b Solder a jumper wire to all pins on the other side of the Berg strip.

c Solder two resistors to the Berg strip, one at each end between the end

pins.

d Solder the center of the BNC connector to the center pin of one row on

the Berg strip.

e Solder the ground tab of the BNC connector to the center pin of the

other row on the Berg strip.

Materials Required · Jumper wire

Page 22 of the Agilent Technologies 1690 user manual — To make the test connectors
Page 22 · click to enlarge
Page 22
Chapter 3Testing Performance To make the test connectors

f On two of the test connectors, solder a 20:1 probe. The probe ground goes to the same row of pins on the test connector as the BNC ground tab.

Agilent Technologies 1690 — To make the test connectors diagram (page 22)
Fig. 6 · To make the test connectors

2 Build one test connector using a BNC connector and a 17-by-2 section of Berg strip:

a Solder a jumper wire to all pins on one side of the Berg strip.

b Solder a jumper wire to all pins on the other side of the Berg strip.

c Solder the center of the BNC connector to the center pin of one row on

the Berg strip.

d Solder the ground tab of the BNC connector to the center pin of the

other row on the Berg strip.

Agilent Technologies 1690 — To make the test connectors diagram (page 22)
Fig. 7 · To make the test connectors
Page 23 of the Agilent Technologies 1690 user manual — Set up the test equipment
Page 23 · click to enlarge
Page 23
Chapter 3Testing Performance To set up the test equipment and the logic analyzer

To set up the test equipment and the logic analyzer

Before testing the specifications of the Agilent 1680A,AD-series or 1690A,AD- series logic analyzer, the test equipment and the logic analyzer must be set up and configured.

These instructions include detailed steps for initially setting up the required test equipment and the logic analyzer. Before performing any or all of the tests in this chapter, the following steps must be done.

Equipment Critical Specifications Recommended HP/Agilent Model/Part

Pulse Generator 200 Mhz, 2.5 ns pulse width,

Digitizing Oscilloscope ≥ 6 GHz bandwidth, < 58 ps rise time 54750A w/ 54751A

Digital Multimeter 0.1 mV resolution, 0.005% accuracy 3458A

Function Generator DC offset voltage ±1.5 V 3325B Option 002

Set up the test equipment

1 Turn on the required test equipment listed in the table above. Let them warm up for 30 minutes before beginning any test.

2 Set up the pulse generator according to the following table.

Timebase Channel 2 Trigger Channel 1

DivideDivide ÷ 2 Ampl: 0.50 V Offs: 0.00 V
ModeInt Period: 5.000 ns

Mode: Pulse Divide: Pulse ÷ 2 Width: 2.500 ns Ampl: 0.80 V Offs: -1.30 V COMP: Disabled (LED Off)

ModeSquare Delay: 0.0 ps Ampl: 0.80 V Offs: -1.30 V COMP: Disabled (LED Off)

3 Set up the oscilloscope:

a Select Setup, then select Default Setup.

Equipment Required · Pulse Generator Setup

Page 24 of the Agilent Technologies 1690 user manual — Set up the 1680A,AD-series logic analyzer
Page 24 · click to enlarge
Page 24
Chapter 3Testing Performance To set up the test equipment and the logic analyzer

b Configure the oscilloscope according to the following table.

Acquisition Display Trigger [Shift] ∆ Time

AveragingOn # of averages: 16
Graticule graphs2 Level: 0.0 V Stop src: channel 2 [Enter]

Channel 1 Channel 2 Define meas

Thresholdsuser-defined Units: Volts Upper: - 980 mV Middle: -1.30 V Lower: -1.62 V
Attenuation20.00:1 Scale: 200 mV/div Offset: - 1.300 V
Attenuation20.00:1 Scale: 200 mV/div Offset: - 1.300 V

Set up the 1680A,AD-series logic analyzer

Power-up self tests are done on the logic analyzer system components when power is applied. Any problems reported by the logic analyzer during boot must be cleared before going further. For more information, refer to Chapter 5 and Chapter 8.

1 Turn on the logic analyzer:

a Connect a keyboard and mouse to the rear panel of the logic analyzer.

b Plug in a power cord to the rear panel power connector of the logic

analyzer.

c Turn on the power switch on the logic analyzer front panel.

2 Set up the logic analyzer:

a Wait for the logic analyzer boot up to complete.

b On the logic analyzer desktop, double-click the Agilent Logic Analyzer

icon to launch the application.

Oscilloscope Setup · External Scale · External Scale

Page 25 of the Agilent Technologies 1690 user manual — To perform the logic analyzer self-tests
Page 25 · click to enlarge
Page 25
Chapter 3Testing Performance To set up the test equipment and the logic analyzer

Set up the 1690A,AD-series logic analyzer

Power-up self tests are done on the logic analyzer system components when power is applied. Logic analyzer peripheral communication tests are done when the host PC recognizes the hosted logic analyzer hardware. Any problems reported should be cleared before going further. For more information, refer to Chapter 5 and Chapter 8.

1 Connect the logic analyzer to the host PC.

2 Set up the logic analyzer:

a Wait for the logic analyzer power-up to complete.

b On the host PC desktop, double-click the Agilent Logic Analyzer icon to

launch the application.

c In the Agilent Logic Analyzer application window, ensure the

application reports “Online.”

To perform the logic analyzer self-tests

The Self Test menu checks the major hardware functions of the logic analyzer to verify that it is working correctly. Self-tests can be performed all at once or one at a time. While testing the performance of the logic analyzer, run the self-tests all at once. Refer to Chapter 8 for more information on the logic analyzer self-tests.

1 In the Agilent Logic Analyzer application, choose Help>Self Test... from the main menu.

Page 26 of the Agilent Technologies 1690 user manual — Set up the 1690A,AD-series logic analyzer
Page 26 · click to enlarge
Page 26
Chapter 3Testing Performance To set up the test equipment and the logic analyzer
Agilent Technologies 1690 — Set up the 1690A,AD-series logic analyzer diagram (page 26)
Fig. 8 · Set up the 1690A,AD-series logic analyzer

2 In the Analysis System Self Tests dialog, select the self test options:

  • Include interactive tests — causes interactive tests to appear in the selection lists.
  • Run repetitively — runs the selected tests repetitively until you click Stop.
  • Stop on fail — if you are running multiple tests or running tests repetitively, this causes the tests to stop if there is a failure.
  • Double-click item to start — lets you double-click a test to start it.

3 Set the reporting level.

Higher levels produce increasingly verbose output.

4 Select the tests you want to run.

5 Click Start.

As the tests are running, the results are reported in the lower part of the dialog and saved to a log file.

Page 27 of the Agilent Technologies 1690 user manual — Set up the 1690A,AD-series logic analyzer
Page 27 · click to enlarge
Page 27
Chapter 3Testing Performance To set up the test equipment and the logic analyzer
Agilent Technologies 1690 — Set up the 1690A,AD-series logic analyzer diagram (page 27)
Fig. 9 · Set up the 1690A,AD-series logic analyzer

To stop running test, click Stop.

To reset the self-test options, click Reset.

To view the log file, click Logs..., select the log file you want to view, and click Open.

If, after completing the self tests, you have failures or you have questions about the performance of the logic analysis system, contact Agilent Technologies sales or support at http://www.agilent.com/find/contactus

6 Click Close to close the Analysis System Self Tests dialog.

Page 28 of the Agilent Technologies 1690 user manual — Set up the equipment
Page 28 · click to enlarge
Page 28
Chapter 3Testing Performance To test the threshold accuracy

To test the threshold accuracy

Testing the threshold accuracy verifies the performance of the following specification:

  • Clock and data channel threshold accuracy.

These instructions include detailed steps for testing the threshold settings of pod 1. After testing pod 1, connect and test the rest of the pods one at a time. To test the next pod, follow the detailed steps for pod 1, substituting the next pod for pod 1 in the instructions.

Each threshold test tells you to record a pass/fail reading in the performance test record located at the end of this chapter.

Equipment Critical Specifications Recommended Model/Part

Digital Multimeter 0.1 mV resolution, 0.005% accuracy 3458A

Function Generator Accuracy (5)(10-6) frequency, DC offset voltage ±1.5 V

33250A

BNC Test Connector, 17x2

Set up the equipment

1 If you have not already done so, do the procedure “To set up the test equipment and the logic analyzer” on page 23.

2 Set up the DC source to deliver a DC voltage on the output:

a In the function generator Utility menu, activate the DC Level. All AC

voltage functions will be disabled.

b Enable the high impedance load under the Output Setup menu.

3 Using a BNC-banana cable, connect the voltmeter to one side of the BNC Tee.

4 Connect the BNC Tee to the output of the DC source. Set up the logic analyzer.

Show 4 more labels / short text on this page
Equipment Required · BNC-Banana Cable 11001-60001 · BNC Tee 1250-0781 · BNC Cable 8120-1840
Page 29 of the Agilent Technologies 1690 user manual — Connect and configure the logic analyzer
Page 29 · click to enlarge
Page 29
Chapter 3Testing Performance

To test the threshold accuracy

Connect and configure the logic analyzer

1 Using the 17-by-2 test connector, BNC cable, and probe tip assembly, connect the data and clock channels of Pod 1 to the free side of the BNC Tee.

Agilent Technologies 1690 — Connect and configure the logic analyzer diagram (page 29)
Fig. 10 · Connect and configure the logic analyzer

2 Configure the logic analyzer:

a Click the Bus/Signal Setup icon. The Analyzer Setup dialog opens.

b In the Buses/Signals tab, click Delete All at the bottom of the dialog.

c Using the mouse, activate all Pod 1 channels. Assign channels to bus/

signal name My Bus 1.

Agilent Technologies 1690 — Connect and configure the logic analyzer diagram (page 29)
Fig. 11 · Connect and configure the logic analyzer

d Scroll the channel assignments to the left. Assign the clock/data

channel for the Pod 1 (that is, C1) to My Bus 1.

Agilent Technologies 1690 — Connect and configure the logic analyzer diagram (page 29)
Fig. 12 · Connect and configure the logic analyzer

3 Activate the DC source output.

Page 30 of the Agilent Technologies 1690 user manual — Test the ECL Threshold
Page 30 · click to enlarge
Page 30
Chapter 3Testing Performance To test the threshold accuracy

Test the ECL Threshold

1 Set up the logic analyzer:

a In the Analyzer Setup dialog, click the threshold field for Pod 1. The

Threshold Settings dialog appears.

b In the Threshold Settings dialog, select Standard and ECL (–1.30 V).

Agilent Technologies 1690 — Test the ECL Threshold diagram (page 30)
Fig. 13 · Test the ECL Threshold

c Click OK to close the Threshold Settings dialog.

d Click OK to close the Analyzer Setup dialog.

2 Test the high-to-low transition:

a On the DC source, enter a voltage setting of –1.384 V.

b On the logic analyzer, click the Run icon. The display should show all channels at a logic "0".

Agilent Technologies 1690 — Test the ECL Threshold diagram (page 30)
Fig. 14 · Test the ECL Threshold

3 Test the low-to-high transition:

a On the DC source, enter a voltage setting of –1.216 V.

b On the logic analyzer, click the Run icon. The display should show all channels at a logic "1" (0x1FFFF).

Agilent Technologies 1690 — Test the ECL Threshold diagram (page 30)
Fig. 15 · Test the ECL Threshold

4 Record a PASS/FAIL in the performance test record for Threshold Accuracy Pod 1 - ECL.

Page 31 of the Agilent Technologies 1690 user manual — Test the 0 V User Threshold
Page 31 · click to enlarge
Page 31
Chapter 3Testing Performance

To test the threshold accuracy

Test the 0 V User Threshold

1 Set up the logic analyzer:

a On the logic analyzer, click the Bus/Signal Setup icon. The Analyzer Setup dialog opens.

b In the Analyzer Setup dialog, click the threshold field for Pod 1. The

Threshold Settings dialog appears.

c In the Threshold Settings dialog, select User Defined and enter 0 V in

the associated field.

Agilent Technologies 1690 — Test the 0 V User Threshold diagram (page 31)
Fig. 16 · Test the 0 V User Threshold

d Click OK to close the Threshold Settings dialog.

e Click OK to close the Analyzer Setup dialog.

2 Test the high-to-low transition:

a On the DC source, enter a voltage setting of –0.064 V.

b On the logic analyzer, click the Run icon. The display should show all channels at a logic "0".

3 Test the low-to-high transition:

a On the DC source, enter a voltage setting of 0.064 V.

b On the logic analyzer, click the Run icon. The display should show all channels at a logic "1" (0x1FFFF).

4 Record a PASS/FAIL in the performance test record for Threshold Accuracy Pod 1 - User 0 V.

Page 32 of the Agilent Technologies 1690 user manual — Test the next pod
Page 32 · click to enlarge
Page 32
Chapter 3Testing Performance To test the threshold accuracy

Test the next pod

1 Using the 17-by-2 test connector and probe tip assembly, connect the data and clock channels of the next pod to the output of the function generator until all pods have been tested.

2 Start with “Connect and configure the logic analyzer” on page 29 substituting the next pod to be tested for pod 1.

Page 33 of the Agilent Technologies 1690 user manual — To set up the logic analyzer for the state mode tests
Page 33 · click to enlarge
Page 33
Chapter 3Testing Performance To set up the logic analyzer for the state mode tests

To set up the logic analyzer for the state mode tests

1 Set up the logic analyzer:

a If you have not already done so, do the procedure “To set up the test

equipment and the logic analyzer” on page 23.

b Exit and restart the Agilent Logic Analyzer application to reinitialize

the logic analyzer.

2 Configure the Analyzer Setup dialog:

a Click the Sampling Setup icon.

b Select State - Synchronous Sampling.

c Configure Trigger Position - 100% poststore.

d Select an Acquisition Depth of 8K.

Agilent Technologies 1690 — To set up the logic analyzer for the state mode tests diagram (page 33)
Fig. 17 · To set up the logic analyzer for the state mode test…

e Click the Buses/Signals tab.

f Click the threshold field of one of the pods. The Threshold Settings dialog appears.

g In the Threshold Settings dialog, select Standard and ECL (–1.30 V).

h Click OK to close the Threshold Settings dialog.

i Click OK to close the the Analyzer Setup dialog.

Page 34 of the Agilent Technologies 1690 user manual — 1680A,AD, 1690A,AD - "AA"
Page 34 · click to enlarge
Page 34
Chapter 3Testing Performance To set up the logic analyzer for the state mode tests

3 Configure the trigger according to your logic analyzer:

a In the Listing window, click on the trigger pattern field for My Bus 1 to

select.

b Enter the following pattern for your logic analyzer.

1680A,AD, 1690A,AD - "AA"

1681A,AD, 1691A,AD - "2A"

1682A,AD, 1692A,AD - "AA"

1683A,AD, 1693A,AD - "A"

Agilent Technologies 1690 — 1680A,AD, 1690A,AD - "AA" diagram (page 34)
Fig. 18

c Click the Trigger Setup icon.

d For the Default Storage, select Store Nothing.

Agilent Technologies 1690 — 1680A,AD, 1690A,AD - "AA" diagram (page 34)
Fig. 19

e Click OK to close the Advanced Trigger dialog.

4 Activate the pulse generator data and clock outputs:

a On the pulse generator, enable the channel 1 OUTPUT, channel 1

OUTPUT, channel 2 OUTPUT and channel 2 OUTPUT (LEDs off)

b On the pulse generator, enable the trigger OUTPUT. (LED off)

5 Set up the Markers:

The following procedure is done after the first run of test data is acquired (during one of the state clock mode tests).

a From the main menu, choose Markers>Properties

b In the Marker Properties tab of the Listing Viewer Properties dialog,

Page 35 of the Agilent Technologies 1690 user manual — 1680A,AD, 1690A,AD - "AA"
Page 35 · click to enlarge
Page 35
Chapter 3Testing Performance To set up the logic analyzer for the state mode tests

select the M1 marker.

Agilent Technologies 1690 — To set up the logic analyzer for the state mode tests diagram (page 35)
Fig. 20 · To set up the logic analyzer for the state mode test…

c In the Position box, select Value.

d Click on the Occurs... button and the Value dialog appears.

e Click on the Find occurrences field, and enter 4096.

f Click on the pattern field, then enter the following pattern according to the logic analyzer being tested:

1680A,AD, 1690A,AD - "AA"

1681A,AD, 1691A,AD - "2A"

1682A,AD, 1692A,AD - "AA"

1683A,AD, 1693A,AD - "A"

Agilent Technologies 1690 — 1680A,AD, 1690A,AD - "AA" diagram (page 35)
Fig. 21

g Click OK to close the Value dialog.

Page 36 of the Agilent Technologies 1690 user manual — 1680A,AD, 1690A,AD - "55"
Page 36 · click to enlarge
Page 36
Chapter 3Testing Performance To set up the logic analyzer for the state mode tests

h Repeat steps b through f to configure marker M2 using the following

pattern according to the logic analyzer being tested.

1680A,AD, 1690A,AD - "55"

1681A,AD, 1691A,AD - "15"

1682A,AD, 1692A,AD - "55"

1683A,AD, 1693A,AD - "5"

Agilent Technologies 1690 — 1680A,AD, 1690A,AD - "55" diagram (page 36)
Fig. 22

i Click OK to close the Value dialog.

j In the Listing Viewer Properties dialog, select Beginning Of Data in the “from” field.

Agilent Technologies 1690 — 1680A,AD, 1690A,AD - "55" diagram (page 36)
Fig. 23

The logic analyzer markers are now configured to verify the test data. If the error message "can't find 4096 occurence(s)" does not appear, the test passes. Click OK to close the Listing ViewerProperties dialog

Agilent Technologies 1690 — 1680A,AD, 1690A,AD - "55" diagram (page 36)
Fig. 24

When the above error message appears, one or more samples of test data is incorrect. When this happens, check the following and attempt the test again:

  • All cables are properly connected.
  • Configuration of each test equipment is correct.
  • Logic analyzer is properly set up according to the test procedure.
Page 37 of the Agilent Technologies 1690 user manual — Connect and configure the logic analyzer
Page 37 · click to enlarge
Page 37
Chapter 3Testing Performance To test the single-clock, single-edge, state acquisition

To test the single-clock, single-edge, state acquisition

Testing the single-clock, single-edge, state acquisition verifies the performance of the following specifications:

  • Minimum master-to-master clock time.
  • Maximum state acquisition speed.
  • Setup/Hold time for single-clock, single-edge, state acquisition.

This test checks two combinations of data channels using a single-edge clock at two selected setup/hold times.

Equipment Critical Specifications Recommended Model/Part

Pulse Generator 200 MHz 2.5 ns pulse width,

Digitizing Oscilloscope ≥6 GHz bandwidth,

SMA Coax Cable (Qty 3) 18 GHz bandwidth 8120-4948

BNC Test Connector, 6x2 (Qty 4)

Set up the equipment

If you have not already done so, do the following procedures:

“To set up the test equipment and the logic analyzer” on page 23.

“To set up the logic analyzer for the state mode tests” on page 33.

Connect and configure the logic analyzer

1 Using the 6-by-2 test connectors, connect the first combination of logic analyzer clock and data channels listed in one of the following tables to the pulse generator.

If you are testing a 1680/81/90/91A,AD, you will repeat this test for the second combination.

Equipment Required · Adapter SMA(m)-BNC(f) 1250-1200 · Coupler BNC(m)(m) 1250-0216

Page 38 of the Agilent Technologies 1690 user manual — To test the single-clock, single-edge, state acquisition
Page 38 · click to enlarge
Page 38
Chapter 3Testing Performance To test the single-clock, single-edge, state acquisition

2 Using SMA cables, connect the oscilloscope to the pulse generator channel 1 Output, channel 2 Output, and Trigger Output.

Connect the 1680/81/90/91A,AD Logic Analyzer to the Pulse Generator

Connect to 8133A Channel 2 Output

Connect to 8133A Channel 2 Output

Connect to 8133A Channel 1 Output

1 Pod 1, channel 3 Pod 3, channel 3 Pod 5, channel 3 Pod 7, channel 3

Pod 2, channel 3 Pod 4, channel 3 Pod 6, channel 3 Pod 8, channel 3 *

Pod 1 clock/data channel (Clk 1)

2 Pod 1, channel 11 Pod 3, channel 11 Pod 5, channel 11 Pod 7, channel 11

Pod 2, channel 11 Pod 4, channel 11 Pod 6, channel 11 Pod 8, channel 11*

Pod 1 clock/data channel (Clk 1)

*1680A,AD or 1690A,AD only

Agilent Technologies 1690 — To test the single-clock, single-edge, state acquisition diagram (page 38)
Fig. 25 · To test the single-clock, single-edge, state acquisi…

Testing Combinations

Page 39 of the Agilent Technologies 1690 user manual — To test the single-clock, single-edge, state acquisition
Page 39 · click to enlarge
Page 39
Chapter 3Testing Performance To test the single-clock, single-edge, state acquisition

Connect the 1682/83/92/93A,AD Logic Analyzer to the Pulse Generator

Connect to 8133A Channel 2 Output

Connect to 8133A Channel 2 Output

Connect to 8133A Channel 1 Output

1 Pod 1, channel 3 Pod 2, channel 3 Pod 3, channel 3 Pod 4, channel 3

Pod 1, channel 11 Pod 2, channel 11 Pod 3, channel 11* Pod 4, channel 11*

Pod 1 clock/data channel (Clk 1)

*1682A,AD or 1692A,AD only.

Agilent Technologies 1690 — To test the single-clock, single-edge, state acquisition diagram (page 39)
Fig. 26 · To test the single-clock, single-edge, state acquisi…

3 Activate the data channels that are connected according to one of the previous tables:

a Click on the Bus/Signal Setup icon. The Analyzer Setup dialog opens.

b In the Buses/Signals tab, click Delete All at the bottom of the dialog.

c Using the mouse, activate the data channels being tested. Assign

channels to bus/signal name My Bus 1.

Agilent Technologies 1690 — To test the single-clock, single-edge, state acquisition diagram (page 39)
Fig. 27 · To test the single-clock, single-edge, state acquisi…

d Click OK to close the Analyzer Setup dialog.

Testing Combination

Page 40 of the Agilent Technologies 1690 user manual — Verify the test signal
Page 40 · click to enlarge
Page 40
Chapter 3Testing Performance To test the single-clock, single-edge, state acquisition

Verify the test signal

1 Check the clock period. Using the oscilloscope, verify that the master-to- master clock time is 5.000 ns, +0 ps or −100 ps:

a In the oscilloscope Timebase menu, select Scale: 1.000 ns/div.

b In the oscilloscope Timebase menu, select Position. Using the

oscilloscope knob, position the clock waveform so that a rising edge appears at the left of the display.

c On the oscilloscope, select [Shift] Period: channel 2, then select [Enter]

to display the clock period (Period(2)). If the period is not less than 5.000 ns, go to step d. If the period is less than 5.000 ns, go to step 2.

d In the oscilloscope Timebase menu, increase Position 5.000 ns. If the

period is not less than 5.000 ns, decrease the pulse generator Period in until one of the two periods measured is less than 5.000 ns.

Agilent Technologies 1690 — Verify the test signal diagram (page 40)
Fig. 28 · Verify the test signal

2 Check the data pulse width. Using the oscilloscope, verify that the data pulse width is 2.500 ns, +0 ps or −100 ps.

a In the oscilloscope Timebase menu, select Position. Using the

oscilloscope knob, position the data waveform so that the waveform is centered on the screen.

b On the oscilloscope, select [Shift] + width: channel 1, then select

[Enter] to display the data signal pulse width (+ width(1)).

Page 41 of the Agilent Technologies 1690 user manual — Check the setup/hold combination
Page 41 · click to enlarge
Page 41
Chapter 3Testing Performance To test the single-clock, single-edge, state acquisition

c If the pulse width is outside the limits, adjust the pulse generator

channel 2 width until the pulse width is within limits.

Agilent Technologies 1690 — Check the setup/hold combination diagram (page 41)
Fig. 29 · Check the setup/hold combination

Check the setup/hold combination

The following setup/hold combinations will be tested:

Sample Position (in middle of Window)

1 4.50/-2.0 ns 2.5 ns -3.25 ns

2 -2.0/4.50 ns 2.5 ns +3.25 ns

1 Disable the pulse generator channel 1 COMP (with the LED off).

2 Using the Delay mode of the pulse generator channel 1, position the pulses according to the setup time of the setup/hold combination selected, +0.0 ps or −100 ps as measured on the oscilloscope:

a On the Oscilloscope, select [Define meas] Define ∆ Time - Stop edge:

rising, Edge number 2.

b In the oscilloscope timebase menu, select Position. Using the

oscilloscope knob, position the data waveform so the falling edge is near the center of the display.

Show 4 more labels / short text on this page
Setup/Hold Combinations · Test Combination · Setup/Hold Times · Setup/Hold Window
Page 42 of the Agilent Technologies 1690 user manual — Check the setup/hold combination
Page 42 · click to enlarge
Page 42
Chapter 3Testing Performance To test the single-clock, single-edge, state acquisition

c On the oscilloscope, select [Shift] ∆ Time, then select [Enter] to display

the setup time (∆ Time(1)-(2)).

d Adjust the pulse generator channel 1 Delay until the pulses are aligned

according to the setup time of the setup/hold combination selected, +0.0 ps or –100 ps.

Agilent Technologies 1690 — Check the setup/hold combination diagram (page 42)
Fig. 30 · Check the setup/hold combination

3 Select the logic analyzer sample positions:

a Click the Sampling Setup icon. The Analyzer Setup dialog opens with the Sampling tab displayed.

b Click Sample Positions

c In the Sample Positions dialog, drag the blue bar for My Bus 1 to the

sample position of the first setup/hold combination to be tested (or enter the value in the signal fields).

Page 43 of the Agilent Technologies 1690 user manual — Check the setup/hold combination
Page 43 · click to enlarge
Page 43
Chapter 3Testing Performance To test the single-clock, single-edge, state acquisition
Agilent Technologies 1690 — Check the setup/hold combination diagram (page 43)
Fig. 31 · Check the setup/hold combination

d Click OK to close the Sample Positions dialog.

4 Select the clock to be tested:

The following clock configurations will be used in steps 4, 5, and 6.

Agilent Technologies 1690 — Check the setup/hold combination diagram (page 43)
Fig. 32 · Check the setup/hold combination

a In the Analyzer Setup dialog, click on the Sampling tab.

b In the Sampling tab, click the Master button for the first clock to be

tested (Clk 1) and select Rising Edge.

Page 44 of the Agilent Technologies 1690 user manual — Check the setup/hold combination
Page 44 · click to enlarge
Page 44
Chapter 3Testing Performance To test the single-clock, single-edge, state acquisition

c Click the Master buttons for the remaining clocks and select Don't Care

to turn off the other clocks.

Agilent Technologies 1690 — Check the setup/hold combination diagram (page 44)
Fig. 33 · Check the setup/hold combination

d Connect the clock to be tested to the pulse generator channel 1 output.

e Click OK to close the Analyzer Setup dialog.

5 Verify the test data:

a Click the Run icon.

b If you have not already done so, do “Set up the Markers:” on page 34.

c If the "can't find 4096 occurence(s)" message does not appear, the test

passes.

The test passes when the logic analyzer finds all occurances of the patterns programmed into the Markers. If the test passes, record a "Pass" in the performance test record under single-clock single-edge next to the clock and edge being tested.

6 Test the next clock:

a Click on the Sampling Setup icon.

b Disconnect the clock just tested from the pulse generator.

c Repeat steps 4, 5, and 6 for the next clock configuration listed in step 4

until all listed clock combinations have been tested.

7 Enable the pulse generator channel 1 COMP (with the LED on).

8 Using the Delay mode of the pulse generator channel 1, position the pulses according to the setup/hold combination selected, +0.0 ps or –100 ps as measured on the oscilloscope:

a On the Oscilloscope, select [Define meas] Define ∆ Time - Stop edge:

falling.

b On the oscilloscope, select [Shift] ∆ Time. Select Start src: channel 1,

Page 45 of the Agilent Technologies 1690 user manual — Check the setup/hold combination
Page 45 · click to enlarge
Page 45
Chapter 3Testing Performance To test the single-clock, single-edge, state acquisition

then select [Enter] to display the setup time (∆ Time(1)-(2)).

c Adjust the pulse generator channel 1 Delay until the pulses are aligned

according to the setup time of the setup/hold combination selected, +0.0 ps or –100 ps.

Agilent Technologies 1690 — Check the setup/hold combination diagram (page 45)
Fig. 34 · Check the setup/hold combination

9 Select the clock to be tested:

The following clock configurations will be used in steps 9, 10 and 11.

Agilent Technologies 1690 — Check the setup/hold combination diagram (page 45)
Fig. 35 · Check the setup/hold combination

a In the Analyzer Setup dialog, click the Sampling tab.

b In the Sampling tab, click the Master button for the first clock to be

tested (Clk 1) and select Falling Edge.

Page 46 of the Agilent Technologies 1690 user manual — Check the setup/hold combination
Page 46 · click to enlarge
Page 46
Chapter 3Testing Performance To test the single-clock, single-edge, state acquisition

c Click the Master buttons for the remaining clocks and select Don't Care

to turn off the other clocks.

Agilent Technologies 1690 — Check the setup/hold combination diagram (page 46)
Fig. 36 · Check the setup/hold combination

d Connect the clock to be tested to the pulse generator channel 1 output.

e Click OK to close the Analyzer Setup dialog.

10 Verify the test data:

a Click the Run icon.

b If you have not already done so, do “Set up the Markers:” on page 34.

c If the "can't find 4096 occurence(s)" message does not appear, the test

passes.

The test passes when the logic analyzer finds all occurances of the patterns programmed into the Markers. If the test passes, record a "Pass" in the performance test record under single-clock single-edge next to the clock and edge being tested.

11 Test the next clock:

a Click the Sampling Setup icon.

b Disconnect the clock just tested from the pulse generator.

c Repeat steps 9, 10, and 11 for the next clock configuration listed in step

9 until all listed clock combinations have been tested.

12 Test the next setup/hold combination:

a Click the Bus/Signal Setup icon.

b Disconnect the clock just tested from the pulse generator.

c Repeat steps 1 through 12 for the next setup/hold combination listed on

page 41.

Page 47 of the Agilent Technologies 1690 user manual — Test the next channels (1680/81A,AD and 1690/91A,AD)
Page 47 · click to enlarge
Page 47
Chapter 3Testing Performance To test the single-clock, single-edge, state acquisition

Test the next channels (1680/81A,AD and 1690/91A,AD)

Connect the next combination of data channels and clock channels; then, test them.

Start with “Connect and configure the logic analyzer” on page 37, connect the next combination; then, continue through the complete test.

Page 48 of the Agilent Technologies 1690 user manual — Set up the equipment
Page 48 · click to enlarge
Page 48
Chapter 3Testing Performance To test the multiple-clock state acquisition

To test the multiple-clock state acquisition

Testing the multiple-clock, state acquisition verifies the performance of the following specifications:

  • Minimum master-to-master clock time.
  • Maximum state acquisition speed.
  • Setup/Hold time for multiple-clock, state acquisition.

This test checks two combinations of data using multiple clocks at two selected setup/hold times.

Equipment Critical Specifications Recommended Model/Part

Pulse Generator 200 MHz 3.0 ns pulse width, < 600 ps rise time

Digitizing Oscilloscope ≥ 6 GHz bandwidth, < 58 ps rise time 54750A w/ 54751A

SMA Coax Cable (Qty 3) 18 GHz bandwidth 8120-4948

BNC Test Connector, 6x2 (Qty 4)

Set up the equipment

1 If you have not already done so, do the following procedures:

“To set up the test equipment and the logic analyzer” on page 23.

“To set up the logic analyzer for the state mode tests” on page 33.

2 Increase the pulse generator channel 2 width to 3.000 ns.

Show 4 more labels / short text on this page
Equipment Required · 8133A option 003 · Adapter SMA(m)-BNC(f) 1250-1200 · Coupler BNC(m)(m) 1250-0216
Page 49 of the Agilent Technologies 1690 user manual — Connect and configure the logic analyzer
Page 49 · click to enlarge
Page 49
Chapter 3Testing Performance To test the multiple-clock state acquisition

Connect and configure the logic analyzer

1 Using the 6-by-2 test connectors, connect the first combination of logic analyzer clock and data channels listed in one of the following tables to the pulse generator.

If you are testing a 1680/81/90/91A,AD, you will repeat this test for the second combination.

2 Using SMA cables, connect channel 1, channel 2, and trigger of the oscilloscope to the pulse generator.

Connect the 1680/81/90/91A,AD Logic Analyzer to the Pulse Generator

Connect to 8133A Channel 2 Output

Connect to 8133A Channel 2 Output

Connect to 8133A Channel 1 Output

1 Pod 1, channel 3 Pod 3, channel 3 Pod 5, channel 3 Pod 7, channel 3

Pod 2, channel 3 Pod 4, channel 3 Pod 6, channel 3 Pod 8, channel 3 *

Clock/data channel for Pod 1, 2, 3, and 4 (Clk 1, Clk 2, Clk 3, Clk 4)

2 Pod 1, channel 11 Pod 3, channel 11 Pod 5, channel 11 Pod 7, channel 11

Pod 2, channel 11 Pod 4, channel 11 Pod 6, channel 11 Pod 8, channel 11 *

Clock/data channel for Pod 1, 2, 3, and 4 (Clk 1, Clk 2, Clk 3, Clk 4)

*1680A, AD or 1690A, AD only.

Agilent Technologies 1690 — Connect and configure the logic analyzer diagram (page 49)
Fig. 37 · Connect and configure the logic analyzer

Testing Combinations

Page 50 of the Agilent Technologies 1690 user manual — Connect and configure the logic analyzer
Page 50 · click to enlarge
Page 50
Chapter 3Testing Performance To test the multiple-clock state acquisition

Connect the 1682/83/92/93A,AD Logic Analyzer to the Pulse Generator

Connect to 8133A Channel 2 Output

Connect to 8133A Channel 2 Output

Connect to 8133A Channel 1 Output

1 Pod 1, channel 3 Pod 2, channel 3 Pod 3, channel 3 Pod 4, channel 3

Pod 1, channel 11 Pod 2, channel 11 Pod 3, channel 11 * Pod 4, channel 11 *

Clock/data channel for Pod 1, 2, 3, and 4 (Clk 1, Clk 2, Clk 3, Clk 4)

*1682A, AD or 1692A, AD only.

Agilent Technologies 1690 — Connect and configure the logic analyzer diagram (page 50)
Fig. 38 · Connect and configure the logic analyzer

3 Activate the data channels that are connected according to one of the previous tables:

a Click the Bus/Signal Setup icon. The Analyzer Setup dialog opens.

b Under the Buses/Signals tab, click Delete All at the bottom of the

dialog.

c Using the mouse, activate the data channels being tested. Assign

channels to bus/signal name My Bus 1.

d Click OK to close the Analyzer Setup dialog.

Testing Combination

Page 51 of the Agilent Technologies 1690 user manual — Verify the test signal
Page 51 · click to enlarge
Page 51
Chapter 3Testing Performance To test the multiple-clock state acquisition

Verify the test signal

1 Check the clock period. Using the oscilloscope, verify that the master-to- master clock time is 5.000 ns, +0 ps or –100 ps:

a In the oscilloscope Timebase menu, select Scale: 1.000 ns/div.

b In the oscilloscope Timebase menu, select Position. Using the

oscilloscope knob, position the clock waveform so that a rising edge appears at the left of the display.

c On the oscilloscope, select [Shift] Period: channel 2, then select [Enter]

to display the clock period (Period(2)). If the period is not less than 5.000 ns, go to step d. If the period is less than 5.000 ns, go to step 2.

d In the oscilloscope Timebase menu, increase Position 5.000 ns. If the

period is not less than 5.000 ns, decrease the pulse generator Period in 10 ps increments until one of the two periods measured is less than 5.000 ns.

Agilent Technologies 1690 — Verify the test signal diagram (page 51)
Fig. 39 · Verify the test signal

2 Check the data pulse width. Using the oscilloscope verify that the data pulse width is 3.000 ns, +0 ps or −100 ps:

a In the oscilloscope Timebase menu, select Position. Using the

oscilloscope knob, position the data waveform so that the waveform is centered on the screen.

Page 52 of the Agilent Technologies 1690 user manual — Check the setup/hold with single clock edges, multiple clocks
Page 52 · click to enlarge
Page 52
Chapter 3Testing Performance To test the multiple-clock state acquisition

b On the oscilloscope, select [Shift] + width: channel 1, then select

[Enter] to display the data signal pulse width (+ width (1)).

c If the pulse width is outside the limits, adjust the pulse generator

channel 2 width until the pulse width is within limits.

Agilent Technologies 1690 — Check the setup/hold with single clock edges, multiple clocks diagram (page 52)
Fig. 40 · Check the setup/hold with single clock edges, multip…

Check the setup/hold with single clock edges, multiple clocks

The following setup/hold combinations will be tested.

Sample Position (in middle of Window)

1 5.0/-2.0 ns 3.0 ns -3.5 ns

2 -1.5/4.50 ns 3.0 ns +3.0 ns

1 Disable the pulse generator channel 1 COMP (LED off).

2 Using the Delay mode of the pulse generator channel 1, position the pulses according to the setup time of the setup/hold combination selected, +0.0 ps or −100 ps as measured on the oscilloscope:

a On the Oscilloscope, select [Define meas] Define ∆ Time - Stop edge:

rising, Edge number 2.

Show 4 more labels / short text on this page
Setup/Hold Combinations · Test Combination · Setup/Hold Times · Setup/Hold Window
Page 53 of the Agilent Technologies 1690 user manual — Check the setup/hold with single clock edges, multiple clocks
Page 53 · click to enlarge
Page 53
Chapter 3Testing Performance To test the multiple-clock state acquisition

b In the oscilloscope timebase menu, select Position. Using the

oscilloscope knob, position the data waveform so the falling edge is near the center of the display.

c On the oscilloscope, select [Shift] ∆ Time, then select [Enter] to display

the setup time (∆ Time(1)-(2)).

d Adjust the pulse generator channel 1 Delay until the pulses are aligned

according to the setup time of the setup/hold combination selected, +0.0 ps or –100 ps.

Agilent Technologies 1690 — Check the setup/hold with single clock edges, multiple clocks diagram (page 53)
Fig. 41 · Check the setup/hold with single clock edges, multip…

3 Select the clocks to be tested:

a Click the Sampling Setup icon. The Analyzer Setup dialog opens.

b In the Sampling tab, click the Master button for one of the clocks and

select Rising Edge.

c Repeat the above steps for each of the remaining clocks until all clocks

have been configured with Rising Edge.

Agilent Technologies 1690 — Check the setup/hold with single clock edges, multiple clocks diagram (page 53)
Fig. 42 · Check the setup/hold with single clock edges, multip…
Page 54 of the Agilent Technologies 1690 user manual — Check the setup/hold with single clock edges, multiple clocks
Page 54 · click to enlarge
Page 54
Chapter 3Testing Performance To test the multiple-clock state acquisition

d Connect all clock channels to the pulse generator channel 1 output.

e Click OK to close the Analyzer Setup dialog.

4 Select the logic analyzer sample positions:

a Click the Sampling Setup icon. The Analyzer Setup dialog opens with the Sampling tab displayed.

b Click Sample Positions

c In the Sample Positions dialog, drag the blue bar for My Bus 1 to the

sample position of the first setup/hold combination to be tested (or enter the value in the signal fields).

Agilent Technologies 1690 — Check the setup/hold with single clock edges, multiple clocks diagram (page 54)
Fig. 43 · Check the setup/hold with single clock edges, multip…

d Click OK to close the Sample Positions dialog.

5 Verify the test data:

a Click the Run icon.

b If you have not already done so, do “Set up the Markers:” on page 34.

c If the "can't find 4096 occurence(s)" message does not appear, then the

test passes.

The test passes when the logic analyzer finds all occurances of the patterns programmed into the Markers. If the test passes, record a "Pass" in the performance test record under single-clock single-edge next to the clock and edge being tested.

Page 55 of the Agilent Technologies 1690 user manual — Check the setup/hold with single clock edges, multiple clocks
Page 55 · click to enlarge
Page 55
Chapter 3Testing Performance To test the multiple-clock state acquisition

6 Enable the pulse generator channel 1 COMP (with the LED on).

7 Using the Delay mode of the pulse generator channel 1, position the pulses according to the setup/hold combination selected, +0.0 ps or –100 ps:

a On the Oscilloscope, select [Define meas] Define ∆ Time - Stop edge:

falling.

b On the oscilloscope, select [Shift] ∆ Time. Select Start src: channel 1,

then select [Enter] to display the setup time (∆ Time(1)-(2)).

c Adjust the pulse generator channel 1 Delay until the pulses are aligned

according to the setup time of the setup/hold combination selected, +0.0 ps or –100 ps.

Agilent Technologies 1690 — Check the setup/hold with single clock edges, multiple clocks diagram (page 55)
Fig. 44 · Check the setup/hold with single clock edges, multip…

8 Select the clocks to be tested:

a Click the Sampling Setup icon. The Analyzer Setup dialog opens.

b In the Sampling tab, click the Master button for one of the clocks and

select Falling Edge.

c Repeat the above steps for each of the remaining clocks until all clocks

Page 56 of the Agilent Technologies 1690 user manual — Test the next channels (1680/81A, AD and 1690/91A, AD)
Page 56 · click to enlarge
Page 56
Chapter 3Testing Performance To test the multiple-clock state acquisition

have been configured with Falling Edge.

Agilent Technologies 1690 — Test the next channels (1680/81A, AD and 1690/91A, AD) diagram (page 56)
Fig. 45 · Test the next channels (1680/81A, AD and 1690/91A, A…

d Click OK to close the Analyzer Setup dialog.

9 Verify the test data:

a Click the Run icon.

b If you have not already done so, do “Set up the Markers:” on page 34.

c If the "can't find 4096 occurence(s)" message does not appear, the test

passes.

The test passes when the logic analyzer finds all occurances of the patterns programmed into the Markers. If the test passes, record a "Pass" in the performance test record under single-clock single-edge next to the clock and edge being tested.

10 Test the next setup/hold combination:

a Click the Bus/Signal Setup icon.

b Disconnect the clock just tested from the pulse generator.

c Repeat steps 1 through 10 for the next setup/hold combination listed in

step 1 in page 52.

When aligning the data and clock waveforms using the oscilloscope, align the waveforms according to the setup time of the setup/hold combination being tested, +0.0 ps or −100 ps.

Test the next channels (1680/81A, AD and 1690/91A, AD)

Connect the next combination of data channels and clock channels, then repeat the previous test.

Start with “Connect and configure the logic analyzer” on page 49, connect the next combination, then continue through the complete test.

Page 57 of the Agilent Technologies 1690 user manual — Set up the equipment
Page 57 · click to enlarge
Page 57
Chapter 3Testing Performance To test the single-clock, multiple-edge, state acquisition

To test the single-clock, multiple-edge, state acquisition

Testing the single-clock, multiple-edge, state acquisition verifies the performance of the following specifications:

  • Minimum master-to-master clock time.
  • Maximum state acquisition speed.
  • Setup/Hold time for single-clock, multiple-edge, state acquisition.

This test checks two combinations of data using a multiple-edge single clock at two selected setup/hold times.

Equipment Critical Specifications Recommended Model/Part

Pulse Generator 200 MHz 3.0 ns pulse width, < 600 ps rise time 8133A option 003

Digitizing Oscilloscope ≥ 6 GHz bandwidth, < 58 ps rise time 54750A w/ 54751A

SMA Coax Cable (Qty 3) 18 GHz bandwidth 8120-4948

BNC Test Connector, 6x2 (Qty 4)

Set up the equipment

1 If you have not already done so, do the following procedures:

“To set up the test equipment and the logic analyzer” on page 23

“To set up the logic analyzer for the state mode tests” on page 33

2 Modify the following pulse generator settings:

Period10.000 ns
Channel 2Width 3.000 ns
Channel 2Pulse ÷ 1
Channel 1Pulse
Channel 1Width 5.000 ns

Equipment Required · Adapter SMA(m)-BNC(f) 1250-1200 · Coupler BNC(m)(m) 1250-0216

Page 58 of the Agilent Technologies 1690 user manual — Connect and configure the logic analyzer
Page 58 · click to enlarge
Page 58
Chapter 3Testing Performance To test the single-clock, multiple-edge, state acquisition

Connect and configure the logic analyzer

1 Using the 6-by-2 test connectors, connect the first combination of logic analyzer clock and data channels listed in one of the following tables to the pulse generator.

If you are testing a 1680/81/90/91A,AD, you will repeat this test for the second combination.

2 Using the SMA cables, connect channel 1, channel 2, and trigger from the oscilloscope to the pulse generator.

Connect the 1680/81/90/91A,AD Logic Analyzer to the Pulse Generator

Connect to 8133A Channel 2 Output

Connect to 8133A Channel 2 Output Connect to 8133A Channel 1 Output

1 Pod 1, channel 3 Pod 3, channel 3 Pod 5, channel 3 Pod 7, channel 3

Pod 2, channel 3 Pod 4, channel 3 Pod 6, channel 3 Pod 8, channel 3 *

Pod 1 clock/data channel (Clk1)

2 Pod 1, channel 11 Pod 3, channel 11 Pod 5, channel 11 Pod 7, channel 11

Pod 2, channel 11 Pod 4, channel 11 Pod 6, channel 11 Pod 8, channel 11 *

Pod 1 clock/data channel (Clk1)

*1680A,AD or 1690A,AD only.

Agilent Technologies 1690 — Connect and configure the logic analyzer diagram (page 58)
Fig. 46 · Connect and configure the logic analyzer

Testing Combinations

Page 59 of the Agilent Technologies 1690 user manual — Connect and configure the logic analyzer
Page 59 · click to enlarge
Page 59
Chapter 3Testing Performance To test the single-clock, multiple-edge, state acquisition

Connect the 1682/83/92/93A,AD Logic Analyzer to the Pulse Generator

Connect to 8133A Channel 2 Output

Connect to 8133A Channel 2 Output Connect to 8133A Channel 1 Output

1 Pod 1, channel 3 Pod 2, channel 3 Pod 3, channel 3 Pod 4, channel 3

Pod 1, channel 3 Pod 2, channel 3 Pod 3, channel 3 * Pod 4, channel 3 *

Pod 1 clock/data channel (Clk1)

*1682A,AD or 1692A,AD only.

Agilent Technologies 1690 — Connect and configure the logic analyzer diagram (page 59)
Fig. 47 · Connect and configure the logic analyzer

3 Activate the data channels that are connected according to one of the previous tables:

a Click the Bus/Signal Setup icon. The Analyzer Setup dialog opens.

b In the Buses/Signals tab, click Delete All at the bottom of the dialog.

c Using the mouse, activate the data channels being tested. Assign

channels to bus/signal name My Bus 1.

d Click OK to close the Analyzer Setup dialog.

Testing Combination

Page 60 of the Agilent Technologies 1690 user manual — Verify the test signal
Page 60 · click to enlarge
Page 60
Chapter 3Testing Performance To test the single-clock, multiple-edge, state acquisition

Verify the test signal

1 Check the clock period. Using the oscilloscope, verify that the master-to- master clock time is 5.000 ns, +0 ps or –100 ps:

a Enable the pulse generator channel 1, channel 2, and trigger outputs

(LED off).

b In the oscilloscope Timebase menu, select Scale: 2.000 ns/div.

c In the oscilloscope Timebase menu, select Position. Using the

oscilloscope knob, position the clock waveform so that a rising edge appears at the left of the display.

d On the oscilloscope, select [Shift] + width: channel 2, then select

[Enter] to display the master-to-master clock time (+ width(2)). If the positive-going pulse width is more than 5.000 ns, go to step e. If the positive-going pulse width is less than or equal to 5.000 ns but greater than 4.900 ns, go to step 2.

e On the oscilloscope, select [Shift] - width: channel 2, then select [Enter]

(- width(2)). If the negative pulse width is less than or equal to 5.000 ns but greater than 4.900 ns, go to step 2.

f Adjust the pulse generator Period and Channel 1 width until the oscilloscope + width (2) or - width (2) reads less than or equal to 5.000 ns, but greater than 4.900 ns.

Agilent Technologies 1690 — Verify the test signal diagram (page 60)
Fig. 48 · Verify the test signal
Page 61 of the Agilent Technologies 1690 user manual — Check the setup/hold with single clock, multiple clock edges
Page 61 · click to enlarge
Page 61
Chapter 3Testing Performance To test the single-clock, multiple-edge, state acquisition

2 Check the data pulse width. Using the oscilloscope, verify that the data pulse width is 3.000 ns, +0 ps or −100 ps:

a In the oscilloscope Timebase menu, select Scale: 1.000 ns/div.

b In the oscilloscope Timebase menu, select Position. Using the

oscilloscope knob, position the data waveform so that the waveform is centered on the screen.

c On the oscilloscope, select [Shift] + width: channel 1, then select

[Enter] to display the data signal pulse width (+ width(1)).

d If the pulse width is outside the limits, adjust the pulse generator

channel 2 width until the pulse width is within limits.

Agilent Technologies 1690 — Check the setup/hold with single clock, multiple clock edges diagram (page 61)
Fig. 49 · Check the setup/hold with single clock, multiple clo…

Check the setup/hold with single clock, multiple clock edges

The following setup/hold combinations will be tested.

Sample Position (in middle of Window)

1 5.0/-2.0 ns 3.0 ns -3.5 ns

2 -1.5/4.50 ns 3.0 ns +3.0 ns

Show 4 more labels / short text on this page
Setup/Hold Combinations · Test Combination · Setup/Hold Times · Setup/Hold Window
Page 62 of the Agilent Technologies 1690 user manual — Check the setup/hold with single clock, multiple clock edges
Page 62 · click to enlarge
Page 62
Chapter 3Testing Performance To test the single-clock, multiple-edge, state acquisition

1 Using the Delay mode of the pulse generator channel 2, position the pulses according to the setup time of the setup/hold combination selected, +0.0 ps or –100 ps:

a On the Oscilloscope, select [Define meas] Define ∆ Time - Stop edge:

rising.

b In the oscilloscope timebase menu, select Position. Using the

oscilloscope knob, position the falling edge of the data waveform so that it is near the center of the display.

c On the oscilloscope, select [Shift] ∆ Time. Select Start src: channel 1,

then select [Enter] to display the setup time (∆ Time(1)-(2)).

d Adjust the pulse generator channel 2 Delay until the pulses are aligned

according to the setup time of the setup/hold combination selected, +0.0 ps or –100 ps.

Agilent Technologies 1690 — Check the setup/hold with single clock, multiple clock edges diagram (page 62)
Fig. 50 · Check the setup/hold with single clock, multiple clo…

2 Select the logic analyzer sample positions:

a Click the Sampling Setup icon. The Analyzer Setup dialog opens with the Sampling tab displayed.

b Click Sample Positions

c In the Sample Positions dialog, drag the blue bar for My Bus 1 to the

sample position of the first setup/hold combination to be tested (or enter the value in the signal fields).

Page 63 of the Agilent Technologies 1690 user manual — Check the setup/hold with single clock, multiple clock edges
Page 63 · click to enlarge
Page 63
Chapter 3Testing Performance To test the single-clock, multiple-edge, state acquisition

d Click OK to close the Sample Positions dialog.

3 Select the clock to be tested:

The following clock configurations will be used in steps 3, 4, and 5.

Agilent Technologies 1690 — Check the setup/hold with single clock, multiple clock edges diagram (page 63)
Fig. 51 · Check the setup/hold with single clock, multiple clo…

a In the Analyzer Setup dialog, click the Sampling tab.

b In the Sampling tab, click the Master button for the first clock to be

tested (Clk 1) and select Both Edges.

c Click the Master buttons for the remaining clocks and select Don't Care

Page 64 of the Agilent Technologies 1690 user manual — Check the setup/hold with single clock, multiple clock edges
Page 64 · click to enlarge
Page 64
Chapter 3Testing Performance To test the single-clock, multiple-edge, state acquisition

to turn off the other clocks.

Agilent Technologies 1690 — Check the setup/hold with single clock, multiple clock edges diagram (page 64)
Fig. 52 · Check the setup/hold with single clock, multiple clo…

d Connect the clock to be tested to the pulse generator channel 1 output.

e Click OK to close the Analyzer Setup dialog.

4 Verify the test data:

a Click the Run icon.

b If you have not already done so, do “Set up the Markers:” on page 34.

c If the "can't find 4096 occurence(s)" does not appear, then the test

passes.

The test passes when the logic analyzer finds all occurrences of the patterns programmed into the Markers. If the test passes, record a "Pass" in the performance test record under single-clock single-edge next to the clock and edge being tested.

5 Test the next clock:

a Click the Sampling Setup icon.

b Disconnect the clock just tested from the pulse generator.

c Repeat steps 3, 4, and 5 for the next clock configuration listed in step 4

until all listed clock combinations have been tested.

6 Test the next setup/hold combination:

a Click the Bus/Signal Setup icon.

b Disconnect the clock just tested from the pulse generator.

c Repeat steps 1 through 6 for the next setup/hold combination listed in

step 1 in page 61.

Page 65 of the Agilent Technologies 1690 user manual — Test the next channels (1680/81A,AD and 1690/91A,AD)
Page 65 · click to enlarge
Page 65
Chapter 3Testing Performance To test the single-clock, multiple-edge, state acquisition

Test the next channels (1680/81A,AD and 1690/91A,AD)

Connect the next combination of data channels and clock channels, then repeat the previous test.

Start with “Connect and configure the logic analyzer” on page 58, connect the next combination, then continue through the complete test.

Page 66 of the Agilent Technologies 1690 user manual — Set up the equipment
Page 66 · click to enlarge
Page 66
Chapter 3Testing Performance To test the time interval accuracy

To test the time interval accuracy

Testing the time interval accuracy does not check a specification, but does check the following:

  • 125 MHz oscillator

This test verifies that the 125 MHz timing acquisition synchronizing oscillator is operating within limits.

Equipment Critical Specifications Recommended Model/Part

Pulse Generator 200 MHz 2.5 ns pulse width, < 600 ps rise time 8133A option 003

Function Generator Accuracy≤ (5)(10-6) x frequency 33250A

BNC Test Connector, 6x2

Set up the equipment

1 Set up the logic analyzer:

a If you have not already done so, do the procedure “To set up the test

equipment and the logic analyzer” on page 23.

b Exit and restart the Agilent Logic Analyzer applications to reinitialize

the logic analyzer.

2 Set up the pulse generator according to the following table.

Timebase Channel 1 Trigger

ModeExt Mode: Square Divide: Divide ÷ 1
Delay0.000 ns Ampl: 0.50 V
High-0.90 V Ampl: 0.50 V
Low-1.70 V Offs: 0.00V
COMPDisabled (LED off)
Show 4 more labels / short text on this page
Equipment Required · SMA Cable 8120-4948 · Adapter BNC(m)-SMA(f) 1250-2015 · Pulse Generator Setup
Page 67 of the Agilent Technologies 1690 user manual — Connect and configure the logic analyzer
Page 67 · click to enlarge
Page 67
Chapter 3Testing Performance To test the time interval accuracy

3 Set up the function generator according to the following table.

Freq40.000 MHz
Ampl1.00 Vpp
Offset0.0 mV

Connect and configure the logic analyzer

1 Using a 6-by-2 test connector, connect channel 0 of Pod 1 to the pulse generator channel 1 output.

2 Using the SMA cable and the BNC adapter, connect the External Input of the pulse generator to the Main Signal of the function generator.

Agilent Technologies 1690 — Connect and configure the logic analyzer diagram (page 67)
Fig. 53 · Connect and configure the logic analyzer

3 Enable the function generator output and the pulse generator Channel 1 output.

4 Configure the Analyzer Setup dialog:

a Click the Sampling Setup icon.

b In the Analyzer Setup dialog, select Timing - Asynchronous Sampling.

c Configure Trigger Position - 100% poststore.

Function Generator Setup · Modulation Off

Page 68 of the Agilent Technologies 1690 user manual — Connect and configure the logic analyzer
Page 68 · click to enlarge
Page 68
Chapter 3Testing Performance To test the time interval accuracy

d Select an Acquisition Depth of 256K.

Agilent Technologies 1690 — Connect and configure the logic analyzer diagram (page 68)
Fig. 54 · Connect and configure the logic analyzer

5 Configure the logic analyzer channels:

a Click the Buses/Signals tab. In the Buses/Signals tab, click Delete All at

the bottom of the dialog.

b Using the mouse, select Pod 1 channel 0 to activate the channel.

Agilent Technologies 1690 — Connect and configure the logic analyzer diagram (page 68)
Fig. 55 · Connect and configure the logic analyzer

c Click the threshold field for Pod 1. In the Threshold Settings dialog,

select Standard and ECL (–1.30).

d Click OK to close the Threshold Settings dialog.

e Click OK to close the Analyzer Setup dialog.

6 Set up the trigger in the Waveform window:

a Select the Simple Trigger field next to bus/signal name My Bus 1.

Page 69 of the Agilent Technologies 1690 user manual — Acquire and verify the test data
Page 69 · click to enlarge
Page 69
Chapter 3Testing Performance To test the time interval accuracy

b In the pop-up menu, select Rising Edge.

Agilent Technologies 1690 — Acquire and verify the test data diagram (page 69)
Fig. 56 · Acquire and verify the test data

Acquire and verify the test data

1 Click the Run icon to fill acquisition memory.

2 Set up the M1 marker for time interval measurement:

a From the main menu, choose Markers>Properties

b In the Marker Properties tab of the Waveform Properties dialog, select

the M1 marker.

c In the Position box, select Value.

Agilent Technologies 1690 — Acquire and verify the test data diagram (page 69)
Fig. 57 · Acquire and verify the test data

d Click Occurs

e In the Value dialog, enter “1” in the Find occurrences field.

Agilent Technologies 1690 — Acquire and verify the test data diagram (page 69)
Fig. 58 · Acquire and verify the test data
Page 70 of the Agilent Technologies 1690 user manual — Acquire and verify the test data
Page 70 · click to enlarge
Page 70
Chapter 3Testing Performance To test the time interval accuracy

f Click OK to close the Value dialog.

3 Set up the M2 marker for time interval measurement:

a In the Marker Properties tab of the Waveform Properties dialog, select

the M2 marker.

b In the Position box, select Value.

c Click Occurs

d In the Value dialog, enter “16384” in the Find occurrences field.

Agilent Technologies 1690 — Acquire and verify the test data diagram (page 70)
Fig. 59 · Acquire and verify the test data

e Click OK to close the Value dialog.

f In the Position box, select M1 in the “from” field. The Position should now read Value Occurs from M1.

Agilent Technologies 1690 — Acquire and verify the test data diagram (page 70)
Fig. 60 · Acquire and verify the test data

g Click Apply; then, click OK to close the Waveform Properties dialog.

4 An Interval Measurement should already be visible in the Markers Toolbar. If not, choose Markers>New Time Interval Measurement from the main menu; in the Time Interval dialog, select from M1 to M2, and click OK. An M1 to M2 time interval field should now be visible in the Markers Toolbar.

5 Click on the Run-Repetitive icon. Allow the logic analyzer to acquire data for at least 100 runs, as reported at the bottom of the window. Observe the M1 to M2 time interval field in the Markers Toolbar and ensure the time interval field is between 40.95571 and 40.96429 µs during the test.

Page 71 of the Agilent Technologies 1690 user manual — Performance Test Record
Page 71 · click to enlarge
Page 71
Chapter 3Testing Performance

Performance Test Record

Agilent 1680/90-Series Logic Analyzer

Serial No Work Order No

Recommended Test Interval - 2 Years/4000 hours Date

Recommended next testing Temperature

± (65 mV + 1.5% of threshold setting)

Pod 1 ECL, ±84 mV 0 V, ±65 mV

Pod 2 ECL, ±84mV 0 V, ± 65mV

Pod 3 ECL, ±84 mV 0 V, ±65 mV

Pod 4 ECL, ±84 mV 0 V, ±65 mV

Pod 5 ECL, ±84 mV 0 V, ±65 mV

Pod 6 ECL, ±84 mV 0 V, ±65 mV

Pod 7 ECL, ±84 mV 0 V, ±65 mV

Pod 8 ECL, ±84 mV 0 V, ±65 mV

Show 21 more labels / short text on this page
Performance Test Record · Test Settings Results · Self-Tests Pass/Fail · ________ · Threshold Accuracy · Pass/Fail Pass/Fail · ________ ________ · Pass/Fail Pass/Fail · ________ ________ · Pass/Fail Pass/Fail · ________ ________ · Pass/Fail Pass/Fail · ________ ________ · Pass/Fail Pass/Fail · ________ ________ · Pass/Fail Pass/Fail · ________ ________ · Pass/Fail Pass/Fail · ________ ________ · Pass/Fail Pass/Fail · ________ ________
Page 72 of the Agilent Technologies 1690 user manual — Performance Test Record
Page 72 · click to enlarge
Page 72
Chapter 3Testing Performance Performance Test Record

Performance Test Record (continued)

All Pods, Channel 3 Setup/Hold Time 4.5/-2.0 ns Clk 1↑ Clk 2↑ Clk 3↑ Clk 4↑

Clk 1↓ Clk 2↓ Clk 3↓ Clk 4↓

Setup/Hold Time -2.0/4.5 ns Clk 1↑ Clk 2↑ Clk 3↑ Clk 4↑

Clk 1↓ Clk 2↓ Clk 3↓ Clk 4↓

All Pods, Channel 11 Setup/Hold Time 4.5/-2.0 ns Clk 1↑ Clk 2↑ Clk 3↑ Clk 4↑

Clk 1↓ Clk 2↓ Clk 3↓ Clk 4↓

Setup/Hold Time -2.0/4.5 ns Clk 1↑ Clk 2↑ Clk 3↑ Clk 4↑

Clk 1↓ Clk 2↓ Clk 3↓ Clk 4↓

All Pods, Channel 3 Setup/Hold Time 5.0/-2.0 ns Clk 1↑ + Clk 2↑ + Clk 3↑ + Clk 4↑

Clk 1↓ + Clk 2↓ + Clk 3↓ + Clk 4↓

Setup/Hold Time -1.5/4.5 ns Clk 1↑+ Clk 2↑ + Clk 3↑ + Clk 4↑

Clk 1↓ + Clk 2↓ + Clk 3↓ + Clk 4↓

All Pods, Channel 11 Setup/Hold Time 5.0/-2.0 ns Clk 1↑ + Clk 2↑ + Clk 3↑ + Clk 4↑

Clk 1↓+ Clk 2↓ + Clk 3↓ + Clk 4↓

Setup/Hold Time -1.5/4.5 ns Clk 1↑ + Clk 2↑ + Clk 3↑ + Clk 4↑

Clk 1↓ + Clk 2↓ + Clk 3↓ + Clk 4↓

Show 9 more labels / short text on this page
Test Settings Results · Single-Clock, Single- Edge Acquisition · Pass/Fail Pass/Fail · Multiple-Clock, Multiple-Edge Acquisition · Pass/Fail Pass/Fail · ________ · ________ · ________ · ________
Page 73 of the Agilent Technologies 1690 user manual — Performance Test Record
Page 73 · click to enlarge
Page 73
Chapter 3Testing Performance

Performance Test Record (continued)

Disable pulse generator, channel 1 COMP (LED off)

All Pods, Channel 3 Setup/Hold Time 5.0/-2.0 ns Clk 1 ↑↓ Clk 2 ↑↓ Clk 3 ↑↓ Clk 4 ↑↓

Setup/Hold Time -1.5/4.5 ns Clk 1 ↑↓ Clk 2 ↑↓ Clk 3 ↑↓ Clk 4 ↑↓

All Pods, Channel 11 Setup/Hold Time 5.0/-2.0 ns Clk 1 ↑↓ Clk 2 ↑↓ Clk 3 ↑↓ Clk 4 ↑↓

Setup/Hold Time -1.5/4.5 ns Clk 1 ↑↓ Clk 2 ↑↓ Clk 3 ↑↓ Clk 4 ↑↓

40.96 µs 40.95571 µs to 40.96429 µs

Show 6 more labels / short text on this page
Performance Test Record · Test Settings Results · Single-Clock, Multiple-Edge Acquisition · Pass/Fail · Time Interval Accuracy · Expected Limits Measured
Page 74 of the Agilent Technologies 1690 user manual — Performance Test Record
Page 74 · click to enlarge
Page 74
Chapter 3Testing Performance Performance Test Record
Page 75 of the Agilent Technologies 1690 user manual — Calibrating and Adjusting
Page 75 · click to enlarge
Page 75

Calibrating and Adjusting

This chapter gives you instructions for calibrating and adjusting the logic analyzer.

Page 76 of the Agilent Technologies 1690 user manual — Logic analyzer calibration
Page 76 · click to enlarge
Page 76
Chapter 4Calibrating and Adjusting

Logic analyzer calibration

The logic analyzer circuitry of the Agilent 1680/90-series logic analyzers does not require an operational accuracy calibration. To test the logic analyzer circuitry against specifications (full calibration), refer to chapter 3, "Testing Performance."

Page 77 of the Agilent Technologies 1690 user manual — Troubleshooting
Page 77 · click to enlarge
Page 77

Troubleshooting

This chapter helps you troubleshoot the logic analyzer to find defective assemblies.

Page 78 of the Agilent Technologies 1690 user manual — To install the fan guard
Page 78 · click to enlarge
Page 78
Chapter 5Troubleshooting

The troubleshooting consists of flowcharts, self-test instructions, and tests. This information is not intended for component-level repair.

If you suspect a problem, start at the top of the first flowchart. During the troubleshooting instructions, the flowcharts will direct you to perform other tests.

The service strategy for this instrument is the replacement of defective assemblies. This instrument can be returned to Agilent Technologies for all service work, including troubleshooting. Contact your nearest Agilent Technologies Sales Office for more details.

NOTE: If any peripheral hardware or software programs were installed by the user into an Agilent 1680A,AD-series logic analyzer, they must be first uninstalled and removed before doing any troubleshooting. Removing user-installed hardware or software will rule out the possibility they are causing problems and/or conflicts in the logic analyzer operating system or application software. Troubleshooting is best done if the instrument is returned to its hardware and software factory configuration.

To install the fan guard

Installing the fan guard is recommended for any power-on troubleshooting for either the Agilent 1680A,AD-series or 1690A,AD-series.

1 Remove the chassis from the sleeve. Follow the procedure “To remove the chassis from the sleeve” on page 114 .

2 Install the fan guard onto the chassis.

a Position the chassis so the handle side is up.

b Slide the fan guard onto the chassis over the fans.

On an Agilent 1680A,AD-series, a guide hole in the fan guard will slide over the standoff post of the bottom left rear foot (adjacent to the acquisition board BNC connectors).

Page 79 of the Agilent Technologies 1690 user manual — To use the flowcharts
Page 79 · click to enlarge
Page 79
Chapter 5Troubleshooting

c Install the optional screws as shown.

Agilent Technologies 1690 — To use the flowcharts diagram (page 79)
Fig. 61 · To use the flowcharts

3 After the required power-on troubleshooting and repair is complete, reverse the above procedure to remove the fan guard and reassemble the instrument.

To use the flowcharts

Flowcharts are the primary tool used to isolate defective assemblies. The flowcharts refer to other tests to help isolate the trouble. The circled letters on the charts indicate connections with the other flowcharts. Start your troubleshooting at the top of the first flowchart.

Page 80 of the Agilent Technologies 1690 user manual — DONE
Page 80 · click to enlarge
Page 80
Chapter 5Troubleshooting Troubleshooting the Agilent 1680A,AD-series

Troubleshooting the Agilent 1680A,AD-series

START

Apply power. Observe

front panel.

No Does the application software

Did the front panel indicators

6A

Are all instrument power

6B

Do the self test. Refer to Chapter 5.

Does the display light

No Do the self tests pass?

Does the instrument finish

board.

Launch the Agilent Logic Analyzer

application software.

DONE

Show 27 more labels / short text on this page
the instrument · No · flash? · launch? · Yes · Yes · Is the application · No · No · supply fans · Online? · running? · Yes · Yes · No · up? · Yes · Replace the · acquisition · No · booting? · Yes · Yes · Is the problem · Yes · still present? · No
Page 81 of the Agilent Technologies 1690 user manual — Troubleshooting the Agilent 1680A,AD-series
Page 81 · click to enlarge
Page 81
Chapter 5Troubleshooting Troubleshooting the Agilent 1680A,AD-series

Ensure the power cord

is properly connected.

Is the power cord

power cord.

Remove the cover. Ensure all front panel

Suspect the power supply. Perform "To test the power supply voltages" in Chapter 5.

connected.

Do all test points

Ensure all instrument cables are connected

Replace the front panel

Replace the power supply.

panel cables.

circuit board.

and properly seated.

Show 22 more labels / short text on this page
Are the power · No · supply fans · running? · Yes · No · No · Are the instrument · Reconnect the · fans running? · connected? · Yes · Yes · cables are properly · Yes · No · Are all cables · connected? · pass? · No · Yes · Reconnect the front
Page 82 of the Agilent Technologies 1690 user manual — Troubleshooting the Agilent 1680A,AD-series
Page 82 · click to enlarge
Page 82
Chapter 5Troubleshooting Troubleshooting the Agilent 1680A,AD-series

Verify the fan voltage.

Perform "To test the fan voltage" in Chapter 5.

Suspect the power supply.

Are the fan voltages OK?

Perform "To test the power supply voltages"

in Chapter 5.

Do all test points pass?

Replace the defective fan.

Replace the distribution board.

Replace the power supply.

Show 13 more labels / short text on this page
Are the power · No · supply fans · running? · Yes · Yes · Are both instrument · fans stopped? · No · No · Yes · No · Yes
Page 83 of the Agilent Technologies 1690 user manual — Troubleshooting the Agilent 1680A,AD-series
Page 83 · click to enlarge
Page 83
Chapter 5Troubleshooting Troubleshooting the Agilent 1680A,AD-series

Connect an external 800x600 PC monito to the instrument.

Is the external monitor readable?

Ensure PCI video board

is seated in the motherboard connector.

video board.

Reseat the PCI video board in the motherboard connector.

Is the PCI video board seated in the

or cables.

Replace PCI video board with known

good board.

Reconnect an external 800x600 PC monitor to the instrument.

Are all cables connected and

Is the external monitor readable?

Replace the LCD display and inverter.

PCI video board. Replace motherboard.

all video cabling.

Show 19 more labels / short text on this page
No · Does the external · monitor light up? · Yes · No · Possible problem with · motherboard or PCI · Yes · Possible problem with · No · LCD display, inverter, · motherboard? · Yes · Yes · No · seated? · Yes No · Reconnect or reseat · Replace defective
Page 84 of the Agilent Technologies 1690 user manual — Troubleshooting the Agilent 1680A,AD-series
Page 84 · click to enlarge
Page 84
Chapter 5Troubleshooting Troubleshooting the Agilent 1680A,AD-series

Yes Does the blue Agilent Logic Analyzer

No Does the instrument boot

motherboard.

Yes Does an *error message appear

Remove power from instrument, remove cover,

Reapply power, observe if any error

check that all cables

messages appear.

are properly seated.

* Microsoft Windows XP Technical Resources will help identify the cause of specific error

No Does an *error message appear?

messages. The Microsoft Windows XP

Technical Resources are currently at http://www.microsoft.com/windowsxp/pro

Does *error message refer to

Yes Does *error message appear re:

message as a hardware failure.

Replace failed hardware as indicated

Replace the hard disk drive.

No Does "starting Windows..."

in the error message.

Does operating system logon screen

system. See "To recover the operating

system".

Show 21 more labels / short text on this page
Replace the CPU · screen appear? · any further? · No Yes · during POST? · No · Yes · Address the specific · No · hard disk drive? · Master Boot · Record? · Yes · No · appear? · Yes · No No · Reinstall operating · Is logon disabled? · appear? · Yes Yes
Page 85 of the Agilent Technologies 1690 user manual — 6A
Page 85 · click to enlarge
Page 85
Chapter 5Troubleshooting Troubleshooting the Agilent 1680A,AD-series

6A

6B

instrument, remove cover, check internal

the Agilent Logic Analyzer application.

IEEE 1394 cabling.

Does the application software launch?

Are all cables connected and

all internal cabling.

Reseat the PCI IEEE board in ISA slot 1*.

* All PCI boards must be installed in the correct slots. See "To remove

Do error messages appear?

Perform "To recover the operating system"

Apply power to the

the ISA board" in Chapter 6.

instrument.

in Chapter 5.

Observe the status LEDs on the acquisition board. Refer to Acquisition board

status LEDs.

Are any green LEDs blinking?

Possible problem with PCI

IEEE 1394 board. Replace

acquisition board. Replace acquisition bd.

the PCI IEEE 1394 board.

Show 17 more labels / short text on this page
Uninstall, then reinstall · Remove power from · No · No · Reconnect or reseat · seated? · Yes · Yes · Yes · No · No · Is the application · Online? · Yes · Yes · No · Possible problem with
Page 86 of the Agilent Technologies 1690 user manual — Troubleshooting the Agilent 1680A,AD-series
Page 86 · click to enlarge
Page 86
Chapter 5Troubleshooting Troubleshooting the Agilent 1680A,AD-series

Possible problem with logic analyzer cables. Do

the cable test in Chapter

5 on suspect pod.

The logic analyzer board if functioning

properly.

known good one.

Replace defective probe tip assembly.

Swap suspect cable assembly with known

good one.

Replace defective cable.

acquisition board.

Show 14 more labels / short text on this page
Does cable · Yes · test pass? · No · Swap suspect probe · Does cable · Yes · test pass? · No · Does cable · Yes · test pass? · No · Replace logic analyzer
Page 87 of the Agilent Technologies 1690 user manual — To test the power supply voltages
Page 87 · click to enlarge
Page 87
Chapter 5Troubleshooting Troubleshooting the Agilent 1680A,AD-series

To check the power-up tests

The power-up self tests on the 1680A,AD-series logic analyzers is performed by the Microsoft Windows XP Professional operating system. As part of the Windows XP Professional power on self test (POST), the presence of all required system components is verified.

1 Close the Agilent Logic Analyzer application and all other applications running on the logic analyzer.

2 Shut down the instrument.

a Click on the Start button in the task bar, then select Shut Down.

b In the Shut Down window, select Shut Down from the menu, then

select the OK button.

c After the instrument turns off, press the power button to again apply

power.

Monitor the boot dialogue. When the text "Starting Windows . . ." appears at the bottom of the screen, this means required system components have been detected and have passed their power-up self-tests.

To test the power supply voltages

Refer to chapter 6, "Replacing Assemblies," for instructions to remove or replace covers and assemblies.

This procedure will not expose any problems related to load regulation; however, it will show most failure modes to over 95% confidence,

1 Close the Agilent Logic Analyzer application and all other applications running on the logic analyzer.

2 Shut down the instrument.

a Click on the Start button in the task bar, then select Shut Down.

b In the Shut Down window, select Shut Down from the menu, then

select the OK button.

Page 88 of the Agilent Technologies 1690 user manual — To check the power-up tests
Page 88 · click to enlarge
Page 88
Chapter 5Troubleshooting Troubleshooting the Agilent 1680A,AD-series

3 Remove the power supply from the instrument. Refer to “To remove the power supply” in Chapter 6.

4 After removing the power supply, connect a power cord to the power supply and plug the power cord into line power.

5 Using DVM, measure the power supply voltages.

Agilent Technologies 1690 — To check the power-up tests diagram (page 88)
Fig. 62 · To check the power-up tests
Agilent Technologies 1690 — To check the power-up tests diagram (page 88)
Fig. 63 · To check the power-up tests

6 Note problems with the power supply, then unplug the power supply from line power. Return to the flow chart.

Power Supply Voltages

Page 89 of the Agilent Technologies 1690 user manual — To test the LCD display signals
Page 89 · click to enlarge
Page 89
Chapter 5Troubleshooting Troubleshooting the Agilent 1680A,AD-series

To test the LCD display signals

Before attempting to do this procedure, ensure that the video signal cable connected to the PCI video board is properly seated. Attempt to reseat the cable two or three times. If other repairs were done to the instrument, and the video is now no longer operating, it is very likely that the video cable is not properly seated.

Refer to chapter 6, "Replacing Assemblies", for instructions to remove or replace covers and assemblies.

1 Remove the sleeve. Refer to chapter 6 for more information on how to remove the sleeve.

2 Connect a power cord to the instrument and apply power.

3 Using an oscilloscope, probe the following pins of J111 for digital signals.

9, 10, 11, 13, 14, 15

29, 30, 31, 33, 34, 35, 37

01680f01.cdr

2, 4, 5

Page 90 of the Agilent Technologies 1690 user manual — To test disk drive voltages
Page 90 · click to enlarge
Page 90
Chapter 5Troubleshooting Troubleshooting the Agilent 1680A,AD-series

4 Using an oscilloscope, probe pins 39 and 40 of J111 for +3.3Vdc

If +3.3Vdc is present on J111 of pins 39 and 40, and digital signals are present on the video data pins indicated above, then the CPU board video circuit is operating properly.

5 Remove power. Allow time for the capacitors in the power supply to discharge before disconnecting the power supply, doing the repair, and reassembling the instrument.

To test disk drive voltages

The following procedure is a guide to help further identify possible problems with either the flexible disk drive or hard disk drive.

Equipment Critical Specification Recommended Model/Part

Digitizing Oscilloscope > 100 MHz Bandwidth 54600B

1 Close the Agilent Logic Analyzer application and all other applications running on the logic analyzer.

2 Shut down the instrument.

a Click on the Start button in the task bar, then select Shut Down.

b In the Shut Down window, select Shut Down from the menu, then

select the OK button.

3 After the instrument turns off, unplug the instrument and remove the cover.

4 Disconnect the suspect disk drive. Remove the disk drive from the chassis and reconnect the cable.

5 Troubleshoot a hard disk drive.

a Apply power to the instrument.

Equipment Required

Page 91 of the Agilent Technologies 1690 user manual — 1 RESET 27 IORDY 34 PDIAG
Page 91 · click to enlarge
Page 91
Chapter 5Troubleshooting Troubleshooting the Agilent 1680A,AD-series

b As the instrument is booting, probe for digital signals on the hard disk

drive connector according to the following table.

Pin No. Signal Voltage Pin No. Signal Voltage Pin No. Signal Voltage

1 RESET 27 IORDY 34 PDIAG

3 - 18 DATA 28 CSEL 35 DA00

20 KEY 29 DMACK 36 DA02

21 DMARQ 31 INTRQ 37 CS0

23 DIOW 32 IOCS16 38 CS1

25 DIOR 33 DA01 39 DASP

Pins 2, 19, 22, 24, 26, 30, 40, are GROUND Pins 41 and 42 are +5 Vdc

6 Troubleshoot a flexible disk drive.

a Apply power to the instrument.

b After the instrument finished booting, launch the Agilent Logic

Analyzer application.

c Insert a formatted flexible disk in the instrument flexible disk drive.

d Attempt to do a File Save of the Agilent Logic Analyzer default

configuration to the flexible disk drive.

Disk Drive Voltages

Page 92 of the Agilent Technologies 1690 user manual — To verify the CD-ROM
Page 92 · click to enlarge
Page 92
Chapter 5Troubleshooting Troubleshooting the Agilent 1680A,AD-series

e While the instrument is attempting to save the file to flexible disk,

probe for digital signals on the flexible disk drive connector according to the following table.

Pin Signals Pin No. Signal

1 +5 V 2 INDEX

3 +5 V 4 DRIVE SELECT

5 +5 V 6 DISK CHANGE

7 NC 8 READY

9 NC 10 MOTOR ON

11 NC 12 DIRECTION SELECT

13 NC 14 STEP

15 0 V 16 WRITE DATA

17 0 V 18 WRITE GATE

19 0 V 20 TRACK 00

21 O V 22 WRITE PROTECT

23 0 V 24 READ DATA

25 0 V 26 SIDE ONE SELECT

7 Repeat steps 1 through 3 above. After the instrument turns off, unplug the instrument.

8 Replace suspect disk drive if necessary, then reassemble the instrument.

To verify the CD-ROM

The CD-ROM drive itself can be tested using an audio CD. Install CD player-style headphones in the CD-ROM audio output jack. With the instrument powered on, insert an audio CD into the CD-ROM drive. If the CD-ROM is operating properly, it should begin playing the audio CD.

Disk Drive Voltages

Page 93 of the Agilent Technologies 1690 user manual — Problems running the Application Software
Page 93 · click to enlarge
Page 93
Chapter 5Troubleshooting Troubleshooting the Agilent 1680A,AD-series

To recover the operating system

To reinstall the operating system

Reinstalling the operating system erases the entire hard disk drive and reinitializes the hard disk drive to its factory configuration. All user data stored on the hard disk drive will be lost. Reinstalling the operating system is necessary in case any system level files or other components of the operating system become corrupted.

The recovery CD-ROM contains an image of the Windows XP Professional operating system. The CD-ROM contains an install key which recognizes whether or not the system motherboard is an Agilent logic analyzer motherboard. If the system motherboard is an Agilent logic analyzer motherboard, then the install key permits the recovery of the operating system from the CD-ROM to the hard disk drive.

1 Apply power to the instrument.

2 After applying power, insert the recovery CD-ROM in the instrument CD- ROM drive.

If the instrument finishes booting, then user files can likely be archived so they don't become lost.

3 Press the on/off button to turn the instrument off.

4 After a few seconds turn the instrument back on.

5 At the prompt, select "Yes" to reinstall the operating system.

It takes about 1 hour to reinstall the operating system and Agilent Logic Analyzer application software. At the end of the operating system reinstallation, the logic analyzer will be in its factory default operating system configuration.

Problems running the Application Software

If there is a problem while running the Agilent Logic Analyzer application software, then the likely cause would be the application if the error or unusual behavior appeared while configuring a window or analyzing data.

In the event of a problem of the Agilent Logic Analyzer application software, do the following steps:

In case the application software becomes unresponsive, do a Ctrl-Alt-Del and follow the queries to abort the application software. Attempt to restart the application software and do a measurement.

If there are still problems running the application software, then uninstall and reinstall

Page 94 of the Agilent Technologies 1690 user manual — Problems with the Operating System
Page 94 · click to enlarge
Page 94
Chapter 5Troubleshooting Troubleshooting the Agilent 1680A,AD-series

the application software.

Problems with the Operating System

Operating system applies to the Agilent 1680A,AD-series logic analyzers. The likely cause would be the operating system if the error or unusual behavior appeared while doing an operating system task, like printing or configuring the network.

In the event of a problem with the operating system, do the following steps:

If error messages appear, consult the operating system documentation for information related to the errors.

In case the whole system becomes unresponsive, turn the instrument off by pressing the on/off button. If pressing the on/off button does not initiate the power-down routine, then press and hold the on/off button for 5 seconds until the instrument turns off.

Turn on the instrument and reattempt the task. If the whole system again becomes responsive then follow the above procedure To reinstall the operating system.

For a host PC controlling an Agilent 1690A,AD-series hosted logic analyzer, responsibility of diagnosing errors and problems with utilized system services is the user's.

Page 95 of the Agilent Technologies 1690 user manual — DONE
Page 95 · click to enlarge
Page 95
Chapter 5Troubleshooting Troubleshooting the Agilent 1690A,AD-series

Troubleshooting the Agilent 1690A,AD-series

START

Apply power.

Are all instrument and power supply fans

Launch the Agilent Logic Analyzer

on the host PC.

No Does the application software

3A

3B

Do the self test. Refer to Chapter 5.

No Do the self tests pass?

board.

DONE

Show 17 more labels / short text on this page
No · running? · Yes · application software · launch? · Yes · Is the application · No · Online? · Yes · Replace the · acquisition · Yes · Is the problem · Yes · still present? · No
Page 96 of the Agilent Technologies 1690 user manual — Troubleshooting the Agilent 1690A,AD-series
Page 96 · click to enlarge
Page 96
Chapter 5Troubleshooting Troubleshooting the Agilent 1690A,AD-series

Ensure the power cord

is properly connected.

Is the power cord

power cord.

Suspect the power supply.

Perform "To test the power supply voltages"

in Chapter 5.

Do all test points pass?

Ensure all instrument cables are connected

Replace the power supply.

and properly seated.

Show 15 more labels / short text on this page
Are the power · No · supply fans · running? · Yes · No · Reconnect the · connected? · Yes · No · Are the instrument · fans running? · Yes · No · Yes
Page 97 of the Agilent Technologies 1690 user manual — 3A
Page 97 · click to enlarge
Page 97
Chapter 5Troubleshooting Troubleshooting the Agilent 1690A,AD-series

3A

3B

the Agilent Logic Analyzer application.

Is the IEEE 1394 cable connected and

IEEE 1394 cable.

Does the application software launch?

Perform " To verify connectivity"

in Chapter 5.

Do error messages appear?

Possible problem with host operating system.

the Agilent Logic Analyzer application.

Consult host PC documentation.

Remove the cover and observe the status LEDs on the acquisition board.

Are any green LEDs blinking?

Possible problem with PCI IEEE 1394 board on

acquisition board. Replace acquisition bd.

host PC. Replace

the board.

Show 17 more labels / short text on this page
Uninstall, then reinstall · No · Reconnect or reseat · seated? · Yes · No · Yes · Yes · No · Uninstall, then reinstall · Is connectivity · verified? · No · Yes · Yes · No · Possible problem with
Page 98 of the Agilent Technologies 1690 user manual — Troubleshooting the Agilent 1690A,AD-series
Page 98 · click to enlarge
Page 98
Chapter 5Troubleshooting Troubleshooting the Agilent 1690A,AD-series

Possible problem with logic analyzer cables. Do

the cable test in Chapter

5 on suspect pod.

The logic analyzer board if functioning

properly.

known good one.

Replace defective probe tip assembly.

Swap suspect cable assembly with known

good one.

Replace defective cable.

acquisition board.

Show 14 more labels / short text on this page
Yes · Does cable · test pass? · No · Swap suspect probe · Yes · Does cable · test pass? · No · Yes · Does cable · test pass? · No · Replace logic analyzer
Page 99 of the Agilent Technologies 1690 user manual — Task Manager
Page 99 · click to enlarge
Page 99
Chapter 5Troubleshooting Troubleshooting the Agilent 1690A,AD-series

To verify connectivity

Using Windows Device Manager and Task Manager, you can quickly determine if the Agilent Logic Analyzer application software is correctly installed on a host PC.

Task Manager

Use Task Manager to see if the agLogicSvc is running. The agLogicSvc is started when the PC is booted and establishes connection with the 1690A,AD-series hosted logic analyzer when the logic analyzer is connected to the PC. If the agLogicSvc service is not listed, then the host PC will not be able to establish an interface with the hosted logic analyzer.

Agilent Technologies 1690 — Task Manager diagram (page 99)
Fig. 64 · Task Manager

If the agLogicSvc service is not listed, then uninstall and reinstall the Agilent Logic Analyzer application software. If the agLogicSvc service still does not appear, then there is a problem with your Windows XP Professional operating system. Consult the documentation for the operating system to further determine why the service is not being installed and run.

Page 100 of the Agilent Technologies 1690 user manual — To test the power supply voltages
Page 100 · click to enlarge
Page 100
Chapter 5Troubleshooting Troubleshooting the Agilent 1690A,AD-series

Device Manager

Use Device Manager to see if the Agilent Logic Analyzer device has been properly installed. The Device Manager should include an entry "Logic Analyzers" with the device "Agilent Technologies 1680/1690 Series Analyzer".

Agilent Technologies 1690 — Device Manager diagram (page 100)
Fig. 65 · Device Manager

If the logic analyzer device is not listed, then uninstall and reinstall the Agilent Logic Analyzer application software. If the device still does not appear, then there is a problem with the Windows XP Professional operating system. Consult the documentation for the operating system to determine why the device is not being installed and run.

To test the power supply voltages

Refer to chapter 6, "Replacing Assemblies," for instructions to remove or replace covers and assemblies.

This procedure will not expose any problems related to load regulation; however, it will show most failure modes to over 95% confidence,

1 Close the Agilent Logic Analyzer application and all other applications running on the logic analyzer.

2 Shut down the instrument.

Page 101 of the Agilent Technologies 1690 user manual — 1-3 +3.4 V 1,2 +5.1 V
Page 101 · click to enlarge
Page 101
Chapter 5Troubleshooting Troubleshooting the Agilent 1690A,AD-series

a Click on the Start button in the task bar, then select Shut Down.

b In the Shut Down window, select Shut Down from the menu, then

select the OK button.

3 Remove the power supply from the instrument. Refer to “To remove the power supply” in Chapter 6.

4 After removing the poser supply, connect a power cord to the power supply and plug the power cord into line power.

5 Using DVM, measure the power supply voltages.

CN1 CN2

Pin Voltage Pin Voltage

1-3 +3.4 V 1,2 +5.1 V

4,5 COM 3 COM

6,7 +3.4 V 4 -5.2 V

8-10 COM 5,6 COM

7 +12 V

8 +5.1 V

9,10 COM

11,12 -5.2 V

13 COM

14 +12 V

Power Supply Voltages

Page 102 of the Agilent Technologies 1690 user manual — Device Manager
Page 102 · click to enlarge
Page 102
Chapter 5Troubleshooting Troubleshooting the Agilent 1690A,AD-series
Agilent Technologies 1690 — Device Manager diagram (page 102)
Fig. 66 · Device Manager

6 Note problems with the power supply, then unplug the power supply from line power. Return to flow chart.

Page 103 of the Agilent Technologies 1690 user manual — To run the self-tests
Page 103 · click to enlarge
Page 103
Chapter 5Troubleshooting

General Troubleshooting

This section includes troubleshooting procedures that can be done on either the Agilent 1680A,AD- or 1690A,AD-series logic analyzers. Before any of these procedures can be done on an Agilent 1690A,AD-series logic analyzer, the logic analyzer must be connected to a host PC and both the host PC and the logic analyzer must be turned on.

To run the self-tests

The Self Test menu checks the major hardware functions of the logic analyzer to verify that it is working correctly. Self-tests can be performed all at once or one at a time. Refer to Chapter 8 for more information on the logic analyzer self-tests.

1 Do the following steps for an Agilent 1690A,AD-series:

a Connect the logic analyzer to a host PC and apply power.

b Apply power to the host PC and allow the PC to finish booting.

2 For an Agilent 1680A,AD-series logic analyzer, apply power to the instrument and allow it to finish booting.

3 Launch the Agilent Logic Analyzer application. When the application launches, observe the status field and ensure it reads “Online.”

4 In the Agilent Logic Analyzer application, choose Help>Self Test... from the main menu.

General Troubleshooting

Page 104 of the Agilent Technologies 1690 user manual — General Troubleshooting
Page 104 · click to enlarge
Page 104
Chapter 5Troubleshooting General Troubleshooting

5 In the Analysis System Self Tests dialog, select the self test options:

  • Include interactive tests — causes interactive tests to appear in the selection lists.
  • Run repetitively — runs the selected tests repetitively until you click Stop.
  • Stop on fail — if you are running multiple tests or running tests repetitively, this causes the tests to stop if there is a failure.
  • Double-click item to start — lets you double-click a test to start it.

6 Set the reporting level.

Higher levels produce increasingly verbose output.

7 Select the tests you want to run.

8 Click Start.

As the tests are running, the results are reported in the lower part of the dialog and saved to a log file.

Page 105 of the Agilent Technologies 1690 user manual — General Troubleshooting
Page 105 · click to enlarge
Page 105
Chapter 5Troubleshooting

The self tests can be run one at a time by clicking on the self-test of interest. The results of the individual test will be reported under Results. For example, if you select the Serial Memory Test and the highest reporting level, the following results should be reported in the status window:

Agilent Technologies 1690 — General Troubleshooting diagram (page 105)
Fig. 67 · General Troubleshooting

To stop a running test, click Stop.

To reset the self-test options, click Reset.

To view the log file, click Logs..., select the log file you want to view, and click Open.

If, after completing the self tests, you have failures or you have questions about the performance of the logic analysis system, contact Agilent Technologies sales or support at http://www.agilent.com/find/contactus

9 Click Close to close the Analysis System Self Tests dialog.

General Troubleshooting

Page 106 of the Agilent Technologies 1690 user manual — Normal Operation
Page 106 · click to enlarge
Page 106
Chapter 5Troubleshooting General Troubleshooting

Acquisition board status LEDs

The acquisition board has four LEDs located close to its IEEE 1394 port. The LEDs report the status of configuration of both the interface field programmable gate array (FPGA) and the IEEE 1394 link layer on the acquisition board.

Agilent Technologies 1690 — Acquisition board status LEDs diagram (page 106)
Fig. 68 · Acquisition board status LEDs

Green LEDs

The green LEDs display the status of loading of the IEEE 1394 link layer.

When the IEEE 1394 is successfully loaded, the system processor on the CPU Motherboard can communicate with the acquisition board. The system processor can then configure and download information from the acquisition board.

Red LED

The Red LED shows the status of the configuration software load of the interface FPGA in the acquisition board.

When the Red LED is on and steady, this indicates the FPGA configuration software has not been successfully loaded into the FPGA. Failure to load the FPGA configuration can be caused by either a failure of the acquisition board or a failure of the PCI IEEE 1394 board. Failure to load the FPGA configuration can also be caused by a missing or misconfigured agLogicSvc service.

Normal Operation

During normal operation, the red LED is off and all three green LEDs are illuminated.

During power up (1680A,AD-series) or connection (1690A,AD-series), the green LEDs first blink on then off. The acquisition board processor attempts to initialize and load the IEEE 1394 link layer. The individual status LEDs represent the success or failure of three steps needed to initalize, load, and then run the IEEE 1394 link layer. When each of the green LEDs illuminate:

Page 107 of the Agilent Technologies 1690 user manual — To test the logic analyzer probe cables
Page 107 · click to enlarge
Page 107
Chapter 5Troubleshooting

LED #1 (closest to the red LED) - the on-board processor is properly initialized and running, attempting to load the IEEE 1394 link layer configuration

LED #2 - the IEEE 1394 link layer is loaded and configured

LED #3 - the IEEE 1394 is loaded, properly configured, and is running

A blinking green LED signifies a failure of one of the above steps. In this case, the acquisition board must be replaced.

When the IEEE 1394 port on the acquisition board is configured and initialized, the agLogicSvc service running on the Windows XP Professional operating system senses the acquisition board. The agLogicSvc service then loads configuration code into the interface FPGA on the acquisition board. When the FPGA is configured, the red LED is turned off. The acquisition board is then properly configured and initialized both to communicate with the system processor and to acquire data.

To test the logic analyzer probe cables

This test allows you to functionally verify the probe cable and probe tip assembly of any of the logic analyzer pods. Only one probe cable can be tested at a time. Repeat this test for each probe cable to be tested.

Equipment Critical Specification Recommended Model/Part

Pulse Generator 200 MHz, 2.5 ns pulse width,

Adapter (Qty 4) SMA (m) - BNC (f) 1250-1200

Coupler (Qty 4) BNC (m)(m) 1250-0216

6x2 Test Connectors (Qty 4)

1 Turn on the equipment and the logic analyzer.

General Troubleshooting · Equipment Required · 8133A Option 003

Page 108 of the Agilent Technologies 1690 user manual — To test the logic analyzer probe cables
Page 108 · click to enlarge
Page 108
Chapter 5Troubleshooting General Troubleshooting

2 Set up the pulse generator according to the following table.

Timebase Channel 2 Trigger Channel 1

ModeInt Mode: Pulse Divide: Divide 1 Mode: Square
Period20.000 ns Divide: Square ÷ 1 Ampl: 0.50 V Delay: 0.000 ns
Offs0.00 V
Ampl0.80 V Ampl: 0.80 V
Offs-1.30 V Offs: -1.30 V
COMPDisabled (LED Off) COMP: Disabled (LED Off)

3 Using four 6-by-2 test connectors, four BNC Couplers, and four SMA (m) - BNC (f) Adapters, connect the logic analyzer to the pulse generator channel outputs (to make the test connectors, see chapter 3, "Testing Performance"):

a Connect the even-numbered channels of the lower byte of the pod

under test and Clk 1 to the pulse generator channel 1 OUTPUT.

b Connect the odd-numbered channels of the lower byte of the pod under

test to the pulse generator channel 1 OUTPUT.

c Connect the even-numbered channels of the upper byte of the pod

under test to the pulse generator channel 2 OUTPUT.

d Connect the odd-numbered channels of the upper byte of the pod

under test to the pulse generator channel 2 OUTPUT.

4 Configure the Analyzer Setup dialog:

a In the Waveform window, click on the Sampling Setup icon.

b In the Analyzer Setup dialog, select State - Synchronous Sampling.

c Configure Trigger Position - 100% poststore.

d Select an Acquisition Depth of 8K.

Pulse Generator Setup

Page 109 of the Agilent Technologies 1690 user manual — To test the logic analyzer probe cables
Page 109 · click to enlarge
Page 109
Chapter 5Troubleshooting

e Click the Master button for the clock to be tested; then, select Both

Edges.

5 Configure the pod under test:

a In the Analyzer Setup dialog, click the Buses/Signals tab

b Click Delete All at the bottom of the window.

c Using the mouse, activate all channels for the pod under test. Assign

channels to bus/signal name My Bus 1.

d Click on the threshold field for the pod under test. The Threshold

Settings dialog appears.

e In the Threshold Settings dialog, select Standard and ECL (-1.30 V).

f Click OK to close the the Threshold Settings dialog.

g Click OK to close the Analyzer Setup dialog.

6 Verify the data:

a Click the Listing - Listing 1 tab.

b In the Listing window, click the Run icon. The display should look

General Troubleshooting

Page 110 of the Agilent Technologies 1690 user manual — To check the BNC Trigger input/output signals
Page 110 · click to enlarge
Page 110
Chapter 5Troubleshooting General Troubleshooting

similar to the figure below.

Agilent Technologies 1690 — To check the BNC Trigger input/output signals diagram (page 110)
Fig. 69 · To check the BNC Trigger input/output signals

7 Repeat steps 3 through 7 to test other logic analyzer cables.

8 Disconnect the test equipment from the logic analyzer.

9 If the display looks like the figure, the cable passed the test.

If the display does not look similar to the figure, there is a possible problem with the cable or probe tip assembly. Causes for cable test failures include the following:

  • Open channel.
  • Channel shortened to a neighboring channel.
  • Channel shortened to either ground or a supply voltage.

Return to Troubleshooting flowchart 7.

To check the BNC Trigger input/output signals

1 Turn on the equipment and the logic analyzer.

2 Set up the DC source to deliver a DC voltage on the output:

a In the function generator Utility menu, activate the DC Level. All AC

voltage functions will be disabled.

b Enable the high impedance load under the Output Setup menu.

3 Connect the equipment to the logic analyzer:

a On the DC source, enter a voltage setting of 0.000 V.

b Using a BNC cable, connect the output of the DC Source to the logic

Page 111 of the Agilent Technologies 1690 user manual — To check the BNC Trigger input/output signals
Page 111 · click to enlarge
Page 111
Chapter 5Troubleshooting

analyzer Trigger In BNC.

c Using a BNC-banana cable, connect the voltmeter to the logic analyzer

Trigger Out BNC.

The voltmeter will display a voltage approximately 3 Vdc.

4 Configure the external trigger:

a Select the Trigger Setup icon.

b In the Advanced Trigger dialog, for Trigger Sequence Step 1, select

“Arm in from” instead of “Anything” as the event to trigger on.

c Then, select “External trigger” as the source of the arming signal.

Agilent Technologies 1690 — To check the BNC Trigger input/output signals diagram (page 111)
Fig. 70 · To check the BNC Trigger input/output signals

d Click OK to close the Advanced Trigger dialog.

5 Verify the external trigger:

a Select the Run icon.

The logic analyzer will report "Waiting in Trigger Step 1...". The voltmeter will display approximately 0 Vdc.

b On the DC source, enter a voltage setting of 3.000 V.

The voltmeter will display approximately 3 Vdc. The logic analyzer will trigger and display a waveform similar to the following:

Agilent Technologies 1690 — To check the BNC Trigger input/output signals diagram (page 111)
Fig. 71 · To check the BNC Trigger input/output signals

c On the DC source, re-enter the voltage setting of 0.000 V.

6 Disconnect the test equipment from the logic analyzer.

General Troubleshooting

Page 112 of the Agilent Technologies 1690 user manual — To test the auxiliary power
Page 112 · click to enlarge
Page 112
Chapter 5Troubleshooting General Troubleshooting

To test the auxiliary power

The +5 V auxiliary power is protected by a current overload protection device. If the current on pins 1 and 39 exceed 0.33 amps, the circuit will open. When the short is removed, the circuit will reset in approximately 1 minute. There should be +5 V after the 1 minute reset time.

Equipment Critical Specifications Recommended Model/Part

Digital Multimeter 0.1 mV resolution, better than 0.005% accuracy E2373A

  • Use the multimeter to verify the +5 V on pins 1 and 39 of the probe cables.
Agilent Technologies 1690 — To test the auxiliary power diagram (page 112)
Fig. 72 · To test the auxiliary power

Equipment Required

Page 113 of the Agilent Technologies 1690 user manual — Replacing Assemblies
Page 113 · click to enlarge
Page 113

Replacing Assemblies

This chapter contains the instructions for removing and replacing the assemblies of the logic analyzer. Also in this chapter are instructions for returning assemblies.

Page 114 of the Agilent Technologies 1690 user manual — To remove the chassis from the sleeve
Page 114 · click to enlarge
Page 114
Chapter 6Replacing Assemblies 1680A,AD-series disassembly/assembly

1680A,AD-series disassembly/assembly

Prepare the instrument for disassembly

Do this procedure before doing any disassembly procedure on the instrument.

1 Close the Agilent Logic Analyzer application software.

2 Gracefully shut down the operating system and remove power when shutdown is complete.

3 Remove the power cord.

4 Move the instrument to a static safe work environment.

To remove the chassis from the sleeve

Before disassembling the instrument, it must be turned off and placed in a static safe work environment. If you haven't already done so, do the previous procedure "Prepare the instrument for disassembly."

1 Using a Torx T-15 screwdriver, remove the two screws that secure the handle to the side of the instrument and lift off the handle.

2 Using a Torx T-10 screwdriver, remove five screws that secure the logic analyzer cables to the rear panel of the logic analyzer.

3 Disconnect the logic analyzer cables from the rear panel. Remove the logic analyzer cables (and spacers, if installed) from the logic analyzer.

4 Using a Torx T-15 screwdriver, remove the screws connecting the four rear feet to the instrument, one screw per foot. Remove each foot from the rear panel.

5 Using a Torx T-10 screwdriver, remove eleven screws that secure the sleeve to the chassis.

6 With the logic analyzer upright, slide the chassis out of the sleeve.

Page 115 of the Agilent Technologies 1690 user manual — To remove the acquisition board
Page 115 · click to enlarge
Page 115
Chapter 6Replacing Assemblies 1680A,AD-series disassembly/assembly

7 Reverse this procedure to install the chassis into the sleeve.

When reassembling, check the following:

all assemblies are properly installed before installing the chassis into the sleeve.

ensure all exposed cables are dressed properly so the sleeve does not cause any damage to the cables.

Agilent Technologies 1690 — To remove the acquisition board diagram (page 115)
Fig. 73 · To remove the acquisition board

To remove the acquisition board

1 Do the procedure "To remove the chassis from the sleeve".

2 Turn the chassis upside down.

3 Disconnect the IEEE 1394 cable from the acquisition board.

4 Using a Torx T-10 screwdriver, remove five screws that secure the probe shroud to the rear panel of the logic analyzer.

5 Using a Torx T-10 screwdriver, remove one screw that secures the acquisition board to the chassis (center of the acquisition board).

Page 116 of the Agilent Technologies 1690 user manual — To remove the acquisition board
Page 116 · click to enlarge
Page 116
Chapter 6Replacing Assemblies 1680A,AD-series disassembly/assembly

6 Slide the acquisition board out the rear panel of the logic analyzer.

Agilent Technologies 1690 — To remove the acquisition board diagram (page 116)
Fig. 74 · To remove the acquisition board

Steps to remove probe shroud from acquisition board.

7 Using a hex screwdriver, remove two hex nuts from the acquisition board trigger BNC connectors.

8 Using a Torx T-10 screwdriver, remove five screws that secure the probe shroud to the acquisition board.

Agilent Technologies 1690 — To remove the acquisition board diagram (page 116)
Fig. 75 · To remove the acquisition board

Reverse this procedure to install the acquisition board.

If the probe shroud requires replacing, then the probe shroud label (part number 01680-94312) must also be ordered and installed on the replacement probe shroud.

Page 117 of the Agilent Technologies 1690 user manual — To remove the power supply
Page 117 · click to enlarge
Page 117
Chapter 6Replacing Assemblies 1680A,AD-series disassembly/assembly

To remove the power supply

1 Do the procedure "To remove the chassis from the sleeve".

2 Disconnect the 16-pin power supply cable and the power sense cable from the distribution board.

3 Using a Torx T-15 screwdriver, remove four screws that secure the power supply to the chassis.

4 Lift the power supply out of the chassis.

5 Disconnect the 14-pin and the 10-pin power supply cables from the distribution board.

Agilent Technologies 1690 — To remove the power supply diagram (page 117)
Fig. 76 · To remove the power supply
Page 118 of the Agilent Technologies 1690 user manual — To remove the hard disk drive
Page 118 · click to enlarge
Page 118
Chapter 6Replacing Assemblies 1680A,AD-series disassembly/assembly

6 Reverse this procedure to install the power supply

When installing a replacement power supply, transfer the power supply cables and the grommet to the replacement power supply as shown:

Agilent Technologies 1690 — To remove the hard disk drive diagram (page 118)
Fig. 77 · To remove the hard disk drive

To remove the hard disk drive

1 Do the procedure "To remove the chassis from the sleeve".

2 Using a Torx T-15 screwdriver, remove four screws that secure the hard disk drive to the chassis.

3 Lift the hard disk drive out of the chassis.

4 Disconnect both the power cable and the data cable from the hard disk drive.

5 Reverse this procedure to install the hard disk drive.

Agilent Technologies 1690 — To remove the hard disk drive diagram (page 118)
Fig. 78 · To remove the hard disk drive
Page 119 of the Agilent Technologies 1690 user manual — To remove the CD-ROM drive assembly
Page 119 · click to enlarge
Page 119
Chapter 6Replacing Assemblies 1680A,AD-series disassembly/assembly

To remove the CD-ROM drive assembly

1 Do the procedure "To remove the chassis from the sleeve".

2 Disconnect the IEEE 1394 cable from the PCI IEEE 1394 interface board. Disengage the cable from the cable clamps on the top of the CD-ROM drive assembly.

3 Remove the data cable, power cable and audio cable from the CD-ROM drive interface board.

4 Using a Torx T-10 screwdriver, remove five screws that secure the CD- ROM drive assembly to the chassis.

5 Slide the CD-ROM assembly out of the chassis.

Agilent Technologies 1690 — To remove the CD-ROM drive assembly diagram (page 119)
Fig. 79 · To remove the CD-ROM drive assembly

Remove the CD-ROM from the assembly

1 1 Using a Torx T-10 screwdriver, remove two screws that secure the CD- ROM interface board to the CD-ROM drive brackets.

2 Remove the interface board from the CD-ROM drive brackets.

A connector on the back of the CD-ROM interface board will disengage from an interface connector on the rear of the CD-ROM drive.

3 Using a Torx T-10 screwdriver, remove two screws, one on each side of the CD-ROM drive brackets, that secure the top bracket to the bottom bracket. Separate the two brackets.

Page 120 of the Agilent Technologies 1690 user manual — To remove the flexible disk drive
Page 120 · click to enlarge
Page 120
Chapter 6Replacing Assemblies 1680A,AD-series disassembly/assembly

4 Using a Torx T-6 screwdriver, remove three screws that secure the CD- ROM drive to the bottom bracket. Lift the CD-ROM drive out of the bottom bracket.

Agilent Technologies 1690 — To remove the flexible disk drive diagram (page 120)
Fig. 80 · To remove the flexible disk drive

5 Reverse this procedure to reassemble and install the CD-ROM drive. Ensure the motherboard is properly installed before installing the CD- ROM drive.

To remove the flexible disk drive

1 Do the procedure "To remove the chassis from the sleeve".

2 Unplug the flexible disk drive cable from the rear of the flexible disk drive.

Agilent Technologies 1690 — To remove the flexible disk drive diagram (page 120)
Fig. 81 · To remove the flexible disk drive
Agilent Technologies 1690 — To remove the flexible disk drive diagram (page 120)
Fig. 82 · To remove the flexible disk drive
Page 121 of the Agilent Technologies 1690 user manual — To remove the flexible disk drive
Page 121 · click to enlarge
Page 121
Chapter 6Replacing Assemblies 1680A,AD-series disassembly/assembly

3 Using a Torx T-8 screwdriver, remove two screws that secure the flexible disk drive to the chassis.

4 Slide the flexible disk drive out the rear of the front panel and out of the chassis.

5 Reverse this procedure to install the flexible disk drive.

Agilent Technologies 1690 — To remove the flexible disk drive diagram (page 121)
Fig. 83 · To remove the flexible disk drive

When installing a new flexible disk drive, industrial double-sided tape (Agilent 0460-2010 or similar) is required to dress the cable onto the flexible disk drive. Apply the tape as shown.

Agilent Technologies 1690 — To remove the flexible disk drive diagram (page 121)
Fig. 84 · To remove the flexible disk drive
Page 122 of the Agilent Technologies 1690 user manual — To remove the PCI boards
Page 122 · click to enlarge
Page 122
Chapter 6Replacing Assemblies 1680A,AD-series disassembly/assembly

To remove the PCI boards

The following PCI boards are installed on the logic analyzer motherboard:

PCI IEEE 1394 board (slot 2, closest to the CD-ROM drive)

PCI display board (slot 3)

slot covers (slot 3 and slot 4)

1 Do the procedure "To remove the chassis from the sleeve".

2 Disconnect the cable from the board to be removed:

PCI IEEE 1394 boardIEEE 1394 cable
PCI display boardvideo cable

3 Using a Torx T-10 screwdriver, remove one screw that secures each PCI board to the rear panel.

4 Slide each board out the top of the chassis.

If the PCI display board is replaced, check to see if tape is covering the video cable connector on the board. If tape covers the connector, remove the tape before installing the PCI display board onto the motherboard.

Agilent Technologies 1690 — To remove the PCI boards diagram (page 122)
Fig. 85 · To remove the PCI boards

5 Reverse this procedure to install the PCI boards.

Page 123 of the Agilent Technologies 1690 user manual — To remove the motherboard
Page 123 · click to enlarge
Page 123
Chapter 6Replacing Assemblies 1680A,AD-series disassembly/assembly

To remove the motherboard

1 Do the procedure "To remove the chassis from the sleeve".

2 Do the procedure "To remove the CD-ROM drive" .

3 Remove both PCI peripheral boards from the motherboard using the procedure "To remove the PCI boards."

4 Disconnect the following cables from the motherboard:

power switch cable from motherboard J1 pins 6 and 8

hard disk drive cable from connector FDD1

power cable from connector CN1

audio cable from connector CD

Agilent Technologies 1690 — To remove the motherboard diagram (page 123)
Fig. 86 · To remove the motherboard

5 Using a Torx T-10 screwdriver, remove the 3 screws that secure the motherboard back plate to the rear panel.

CPU fan cable

Page 124 of the Agilent Technologies 1690 user manual — Transfer the I/O panel to the replacement board
Page 124 · click to enlarge
Page 124
Chapter 6Replacing Assemblies 1680A,AD-series disassembly/assembly

6 Using a Torx T-10 screwdriver, remove the 6 screws that secure the motherboard to the chassis.

Agilent Technologies 1690 — Transfer the I/O panel to the replacement board diagram (page 124)
Fig. 87 · Transfer the I/O panel to the replacement board

7 Reverse this procedure to install the motherboard.

If the motherboard requires replacing, then do the following steps:

Transfer the I/O panel to the replacement board

1 Using a 3/16 inch hex nut driver, remove 6 hex studs that secure the I/O panel to the parallel port, VGA port and COM1 port.

Agilent Technologies 1690 — Transfer the I/O panel to the replacement board diagram (page 124)
Fig. 88 · Transfer the I/O panel to the replacement board
Page 125 of the Agilent Technologies 1690 user manual — Transfer the I/O panel to the replacement board
Page 125 · click to enlarge
Page 125
Chapter 6Replacing Assemblies 1680A,AD-series disassembly/assembly

3 If the I/O panel requires replacement, then EMI gasket (01680-87102) must be installed into the replacement I/O panel. The gasket is installed as follows:

a Two 80 mm sections are installed, one on each side of the top center

screw hole.

b One 160 mm section is installed centered along the bottom edge of the

I/O panel directly beneath the keyboard, USB, COM1, VGA and audio ports.

Agilent Technologies 1690 — Transfer the I/O panel to the replacement board diagram (page 125)
Fig. 89 · Transfer the I/O panel to the replacement board

4 Install the I/O panel onto the replacement motherboard. (Shown in step 2 on previous page.)

5 Install 6 hex studs onto the motherboard ports and tighten.

Page 126 of the Agilent Technologies 1690 user manual — To remove the front panel assembly
Page 126 · click to enlarge
Page 126
Chapter 6Replacing Assemblies 1680A,AD-series disassembly/assembly

To remove the front panel assembly

1 Do the procedure "To remove the chassis from the sleeve".

2 Using diagonal cutters, cut the tie wrap off the on/off cable.

3 Disconnect the display cable from the PCI display board.

4 Disconnect the following cables from the distribution board:

on/off cable (accessed from the bottom of the chassis)

Agilent Technologies 1690 — To remove the front panel assembly diagram (page 126)
Fig. 90 · To remove the front panel assembly

5 Using a Torx T-15 screwdriver, remove four screws that secure the front panel assembly to the chassis. Lift the front panel assembly away from the chassis.

keyboard cable · inverter cable

Page 127 of the Agilent Technologies 1690 user manual — To remove the front panel assembly
Page 127 · click to enlarge
Page 127
Chapter 6Replacing Assemblies 1680A,AD-series disassembly/assembly
Agilent Technologies 1690 — To remove the front panel assembly diagram (page 127)
Fig. 91 · To remove the front panel assembly

6 Reverse this procedure to install the front panel assembly.

After installing the front panel assembly onto the chassis, loop the on/off cable and dress the cable with a cable tie (Agilent part number 1400-0249 or similar). Dressing the cable with a cable tie ensures the cable will not be caught in the sleeve when the chassis is slid into the sleeve during assembly.

Agilent Technologies 1690 — To remove the front panel assembly diagram (page 127)
Fig. 92 · To remove the front panel assembly
Page 128 of the Agilent Technologies 1690 user manual — To disassemble the front panel assembly
Page 128 · click to enlarge
Page 128
Chapter 6Replacing Assemblies 1680A,AD-series disassembly/assembly

To disassemble the front panel assembly

Agilent Technologies 1690 — To disassemble the front panel assembly diagram (page 128)
Fig. 93 · To disassemble the front panel assembly

Remove the front panel circuit board.

1 Remove each of the knobs from the front panel.

2 Using a Torx T-10 screwdriver, remove five screws that secure the front panel circuit board to the front panel.

3 Lift the front panel circuit board out of the front panel.

4 Lift the elastomeric keypad out of the front panel.

Remove the LCD display.

1 Using a Torx T-10 screwdriver, remove four screws, one in each corner, that secures the LCD display to the front panel.

2 Lift the LCD display out of the front panel.

3 Lift the glass lens out of the front panel assembly.

Remove the on/off switch.

1 Using a Torx T-10 screwdriver, remove two screws that secure the on-off switch circuit board to the front panel.

2 Lift the on/off switch circuit board out of the front panel.

3 Lift the on/off switch out of the front panel.

Page 129 of the Agilent Technologies 1690 user manual — To remove the distribution board
Page 129 · click to enlarge
Page 129
Chapter 6Replacing Assemblies 1680A,AD-series disassembly/assembly

Reverse this procedure to assemble the front panel assembly.

If the front panel requires replacement, a product ID label must also be ordered and applied (refer to chapter 7 for the part number).

To remove the distribution board

1 1 Do the following procedures:

To remove the chassis from the sleeve

To remove the front panel assembly

2 Disconnect all cables from the distribution board as shown.

Agilent Technologies 1690 — To remove the distribution board diagram (page 129)
Fig. 94 · To remove the distribution board

3 Using a Torx T-10 screwdriver, remove six screws that secure the distribution board to the chassis.

4 Lift the distribution board away from the chassis.

Page 130 of the Agilent Technologies 1690 user manual — To remove the inverter board
Page 130 · click to enlarge
Page 130
Chapter 6Replacing Assemblies 1680A,AD-series disassembly/assembly

5 Reverse this procedure to install the distribution board.

Agilent Technologies 1690 — To remove the inverter board diagram (page 130)
Fig. 95 · To remove the inverter board

To remove the inverter board

1 Do the procedure "To remove the chassis from the sleeve".

2 Disconnect both the inverter input and inverter LCD inverter output cables from the inverter board.

3 Using a Torx T-10 screwdriver, remove two screws that secure the inverter board to the chassis.

4 Lift the inverter board out of the chassis.

Page 131 of the Agilent Technologies 1690 user manual — To remove the fans
Page 131 · click to enlarge
Page 131
Chapter 6Replacing Assemblies 1680A,AD-series disassembly/assembly

5 Reverse this procedure to install the inverter board.

Agilent Technologies 1690 — To remove the fans diagram (page 131)
Fig. 96 · To remove the fans

To remove the fans

1 Do the procedure "To remove the chassis from the sleeve".

2 Disconnect the fan cables from the distribution board.

3 Using needle-nosed pliers, remove the plastic push fastener insert from the push fasteners securing one of the fans to the chassis.

4 Disengage the fan from the push fasteners and remove the fan through the bottom of the chassis.

5 If necessary, repeat steps 2 through 4 for the remaining fan.

6 Reverse this procedure to install the fans.

Agilent Technologies 1690 — To remove the fans diagram (page 131)
Fig. 97 · To remove the fans
Page 132 of the Agilent Technologies 1690 user manual — 1680A,AD: 01680-94307
Page 132 · click to enlarge
Page 132
Chapter 6Replacing Assemblies 1680A,AD-series disassembly/assembly

To remove the cable tray

Do the following steps only if the cable tray requires replacement.

If the cable tray requires replacing, then a label must also be ordered and applied to the cable tray depending on the 1680A,AD-series instrument the cable tray is installed on:

1680A,AD: 01680-94307

1681A,AD: 01680-94308

1682A,AD: 01680-94309

1683A,AD: 01680-94310

1 Remove the tilt stand from the bottom front feet.

2 Remove the logic analyzer cables from the cable tray

3 Using a Torx T-20 screwdriver, remove four screws that secure the cable tray to the sleeve.

4 Reverse this procedure to install the cable tray.

Agilent Technologies 1690 — 1680A,AD: 01680-94307 diagram (page 132)
Fig. 98
Page 133 of the Agilent Technologies 1690 user manual — To remove the chassis from the sleeve
Page 133 · click to enlarge
Page 133
Chapter 6Replacing Assemblies 1690A,AD-series disassembly/assembly

1690A,AD-series disassembly/assembly

Prepare the instrument for disassembly

Do this procedure before doing any disassembly procedure on the instrument.

1 Close the Agilent Logic Analyzer application software.

2 Disconnect the logic analyzer from the host PC.

3 Remove power and disconnect the power cord.

4 Move the instrument to a static safe work environment.

To remove the chassis from the sleeve

Before disassembling the instrument, it must be turned off and placed in a static safe work environment. If you haven't already done so, do the previous procedure "Prepare the instrument for disassembly."

1 Using a Torx T-15 screwdriver, remove the two screws that secure the handle to the side of the instrument and lift off the handle.

2 Using a Torx T-10 screwdriver, remove five screws that secure the logic analyzer cables (and spacers, if installed) to the front panel of the logic analyzer.

3 Disconnect the logic analyzer cables from the front panel. Remove the logic analyzer cables (and spacers, if installed) from the logic analyzer.

4 Using a Torx T-10 screwdriver, remove thirteen screws that secure the cover to the chassis.

5 With the logic analyzer upright, slide the chassis out of the cover.

Page 134 of the Agilent Technologies 1690 user manual — To remove the fascia
Page 134 · click to enlarge
Page 134
Chapter 6Replacing Assemblies 1690A,AD-series disassembly/assembly

6 Reverse this procedure to install the chassis into the sleeve.

When reassembling, check the following:

all assemblies are properly installed before installing the chassis into the sleeve.

ensure all exposed cables are dressed properly so the sleeve does not cause any damage to the cables.

Agilent Technologies 1690 — To remove the fascia diagram (page 134)
Fig. 99 · To remove the fascia

To remove the fascia

1 Do the procedure "To remove the chassis from the sleeve".

2 Disconnect the on/off cable from the distribution board and remove the cable from the cable clip.

3 Disengage the four tabs in the inside of the front panel that secure the fascia to the front panel.

Agilent Technologies 1690 — To remove the fascia diagram (page 134)
Fig. 100 · To remove the fascia
Page 135 of the Agilent Technologies 1690 user manual — Steps to remove the on/off switch from the fascia assembly
Page 135 · click to enlarge
Page 135
Chapter 6Replacing Assemblies 1690A,AD-series disassembly/assembly

4 Remove the fascia away from the front panel.

If the fascia requires replacement, a product ID label must also be ordered and applied (refer to chapter 7 for the part number). Also, Pod labels (01680-94313) must be ordered and applied in the same way as on the fascia being replaced.

Agilent Technologies 1690 — Steps to remove the on/off switch from the fascia assembly diagram (page 135)
Fig. 101 · Steps to remove the on/off switch from the fascia as…

Steps to remove the on/off switch from the fascia assembly

1 Using a Torx T-10 screwdriver, remove two screws that secure the on-off switch circuit board to the fascia.

2 Lift the on/off switch circuit board out of the fascia.

3 Lift the on/off switch out of the fascia.

Agilent Technologies 1690 — Steps to remove the on/off switch from the fascia assembly diagram (page 135)
Fig. 102 · Steps to remove the on/off switch from the fascia as…

Reverse this procedure to assemble the fascia.

Page 136 of the Agilent Technologies 1690 user manual — To remove the acquisition board
Page 136 · click to enlarge
Page 136
Chapter 6Replacing Assemblies 1690A,AD-series disassembly/assembly

To remove the acquisition board

1 Do the following procedures:

To remove the chassis from the sleeve

To remove the fascia

2 Using a Torx T-10 screwdriver, remove one screw that secures the acquisition board to the deck.

3 Using a Torx T-10 screwdriver, remove five screws that secure the probe shroud to the logic analyzer front panel.

4 Slide the acquisition board out of the logic analyzer front panel.

Agilent Technologies 1690 — To remove the acquisition board diagram (page 136)
Fig. 103 · To remove the acquisition board

Steps to remove the probe shroud from the acquisition board.

1 Using a hex screwdriver, remove two hex nuts from the acquisition board trigger BNC connectors.

2 Using a Torx T-10 screwdriver, remove five screws that secure the probe shroud to the acquisition board.

Agilent Technologies 1690 — To remove the acquisition board diagram (page 136)
Fig. 104 · To remove the acquisition board

Reverse this procedure to install the acquisition board

If the probe shroud requires replacing, then the probe shroud label (part number 01690-94302) must also be ordered and installed on the replacement probe shroud.

Page 137 of the Agilent Technologies 1690 user manual — To remove the power supply
Page 137 · click to enlarge
Page 137
Chapter 6Replacing Assemblies 1690A,AD-series disassembly/assembly

To remove the deck

1 Using a Torx T-10 screwdriver, remove four screws that secure the deck to the chassis.

2 Tilt the rear of the deck up and out of the chassis.

3 Lift the deck up and away from the chassis.

Agilent Technologies 1690 — To remove the deck diagram (page 137)
Fig. 105 · To remove the deck

To remove the power supply

1 Do the following procedures:

To remove the chassis from the sleeve

To remove the front panel assembly

To remove the deck

2 Disconnect the power supply output cables, the on/off cable and the power sense cable from the distribution board.

3 Disconnect the power supply line input cable from the power supply.

4 Using a Torx T-10 screwdriver, remove four screws that secure the power supply to the bottom of the chassis.

Page 138 of the Agilent Technologies 1690 user manual — To remove the distribution board
Page 138 · click to enlarge
Page 138
Chapter 6Replacing Assemblies 1690A,AD-series disassembly/assembly

5 Lift the power supply out of the chassis.

Agilent Technologies 1690 — To remove the distribution board diagram (page 138)
Fig. 106 · To remove the distribution board

6 Reverse this procedure to install the power supply

When installing a replacement power supply, transfer both the power sense cable and the power supply on/off cable to the replacement power supply.

To remove the distribution board

1 Do the following procedures:

To remove the chassis from the sleeve

To remove the front panel assembly

To remove the deck

2 Disconnect the following cables from the distribution board:

power supply output cables (P1, P2)

power sense cable (P7)

fan cables (P5, P6)

3 Using a Torx T-10 screwdriver, remove three screws that secure the distribution board to the chassis.

on/off cable (P3)

Page 139 of the Agilent Technologies 1690 user manual — To remove the fans
Page 139 · click to enlarge
Page 139
Chapter 6Replacing Assemblies 1690A,AD-series disassembly/assembly

4 Slide the distribution board up approx 1 cm, then lift the board away from the chassis.

5 Reverse this procedure to install the distribution board.

Agilent Technologies 1690 — To remove the fans diagram (page 139)
Fig. 107 · To remove the fans

To remove the fans

1 Do the procedure "To remove the chassis from the sleeve".

2 Disconnect the fan cables from the distribution board and remove the cables from the cable clips.

3 Using needle-nosed pliers, remove the plastic push fastener insert from the push fasteners securing one of the fans to the chassis.

4 Disengage the fan from the push fasteners and remove the fan out the top of the chassis.

5 If necessary, repeat steps 2 through 4 for the remaining fan.

Page 140 of the Agilent Technologies 1690 user manual — To remove the line filter
Page 140 · click to enlarge
Page 140
Chapter 6Replacing Assemblies 1690A,AD-series disassembly/assembly

6 Reverse this procedure to install the fans.

Agilent Technologies 1690 — To remove the line filter diagram (page 140)
Fig. 108 · To remove the line filter

To remove the line filter

1 Do the following procedures:

To remove the chassis from the sleeve

To remove the front panel assembly

To remove the deck

2 Disconnect the power supply line input cable from the power supply.

3 Using a Torx T-10 screwdriver, remove two screws that secure the line filter to the rear of the chassis.

4 Remove the line filter out the rear of the chassis.

Agilent Technologies 1690 — To remove the line filter diagram (page 140)
Fig. 109 · To remove the line filter
Page 141 of the Agilent Technologies 1690 user manual — To remove the front panel and front frame
Page 141 · click to enlarge
Page 141
Chapter 6Replacing Assemblies 1690A,AD-series disassembly/assembly

To remove the front panel and front frame

1 Do the following procedures:

To remove the chassis fron the sleeve

To remove the fascia

To remove the acquisition board

2 Remove the trim strips from the top and sides of the front frame.

3 Using a Torx T-15 screwdriver remove four screws that secure the top of the front panel to the front frame.

4 Using a Torx T-15 screwdriver, remove four screws that secure the bottom of the front panel to the front frame, as shown.

5 Using a Torx T-15 screwdriver, remove four screws that secure the front frame to the deck.

Agilent Technologies 1690 — To remove the front panel and front frame diagram (page 141)
Fig. 110 · To remove the front panel and front frame
Page 142 of the Agilent Technologies 1690 user manual — To remove the front panel and front frame
Page 142 · click to enlarge
Page 142
Chapter 6Replacing Assemblies 1690A,AD-series disassembly/assembly
Page 143 of the Agilent Technologies 1690 user manual — Replaceable Parts
Page 143 · click to enlarge
Page 143

Replaceable Parts

This chapter contains information for identifying and ordering replaceable parts for your logic analyzer.

Page 144 of the Agilent Technologies 1690 user manual — Direct mail order system
Page 144 · click to enlarge
Page 144
Chapter 7Replaceable Parts

Replaceable Parts Ordering

Parts listed

To order a part on the list of replaceable parts, quote the Agilent Technologies part number, indicate the quantity desired, and address the order to the nearest Agilent Technologies Sales Office.

Parts not listed

To order a part not on the list of replaceable parts, include the model number and serial number of the module, a description of the part (including its function), and the number of parts required. Address the order to your nearest Agilent Technologies Sales Office.

Direct mail order system

To order using the direct mail order system, contact your nearest Agilent Technologies Sales Office.

Within the USA, Agilent Technologies can supply parts through a direct mail order system. The advantages to the system are direct ordering and shipment from the Agilent Technologies Part Center in Mountain View, California. There is no maximum or minimum on any mail order. (There is a minimum amount for parts ordered through a local Agilent Technologies Sales Office when the orders require billing and invoicing.) Transportation costs are prepaid (there is a small handling charge for each order) and there are no invoices.

For Agilent Technologies to provide these advantages, a check or money order must accompany each order. Mail order forms and specific ordering information are available through your local Agilent Technologies Sales Office. Addresses and telephone numbers are located in a separate document at the back of the service guide.

Page 145 of the Agilent Technologies 1690 user manual — Replaceable Parts List
Page 145 · click to enlarge
Page 145
Chapter 7Replaceable Parts

Exchange Assemblies

Some assemblies are part of an exchange program with Agilent Technologies.

The exchange program allows you to exchange a faulty assembly with one that has been repaired and performance verified by Agilent Technologies.

After you receive the exchange assembly, return the defective assembly to Agilent Technologies. A United States customer has 30 days to return the defective assembly. If you do not return the defective assembly within the 30 days, Agilent Technologies will charge you an additional amount. This amount is the difference in price between a new assembly and that of the exchange assembly. For orders not originating in the United States, contact your nearest Agilent Technologies Sales Office for information.

See Also "To return assemblies," in chapter 6.

Replaceable Parts List

The replaceable parts list is organized by reference designation and shows exchange assemblies, electrical assemblies, then other parts.

The exploded view does not show all of the parts in the replaceable parts list.

Information included for each part on the list consists of the following:

  • Reference designator
  • Agilent Technologies part number
  • Total quantity included with the instrument (Qty)
  • Description of the part

Reference designators used in the parts list are as follows:

  • A Assembly
  • E Miscellaneous Electrical Part
  • F Fuse
  • H Hardware
  • MP Mechanical Part
  • W Cable
Page 146 of the Agilent Technologies 1690 user manual — Exploded View
Page 146 · click to enlarge
Page 146
Chapter 7Replaceable Parts

Exploded View

Agilent Technologies 1690 — Exploded View diagram (page 146)
Fig. 111 · Exploded View

Exploded view of the Agilent 1680A,AD-series logic analyzer

Page 147 of the Agilent Technologies 1690 user manual — Agilent 1680A,AD-Series Replaceable Parts
Page 147 · click to enlarge
Page 147
Chapter 7Replaceable Parts

Agilent 1680A,AD-Series Replaceable Parts

Ref. Des. Agilent Part Number QTY Description

01680-69507 Acquisition Board 136 Channel x 2 Mbit (1680AD)

01680-69508 Acquisition Board 34 Channel x 2 Mbit (1693AD)

01680-69509 Acquisition Board 68 Channel x 2 Mbit (1692AD)

01680-69510 Acquisition Board 102 Channel x 2 Mbit (1691AD)

01680-69517 Acquisition Board 34 Channel x 512 Kbit (1693A)

01680-69518 Acquisition Board 68 Channel x 512 Kbit (1692A)

01680-69519 Acquisition Board 102 Channel x 512 Kbit (1691A)

01680-69520 Acquisition Board 136 Channel x 512 Kbit (1690A)

Replacement Parts A2 01680-66507 1 Acquisition Board 136 Channel x 2 Mbit (1680AD)

A2 01680-66508 1 Acquisition Board 34 Channel x 2 Mbit (1693AD)

A2 01680-66509 1 Acquisition Board 68 Channel x 2 Mbit (1692AD)

A2 01680-66510 1 Acquisition Board 102 Channel x 2 Mbit (1691AD)

A2 01680-66517 1 Acquisition Board 34 Channel x 512 Kbit (1693A)

A2 01680-66518 1 Acquisition Board 68 Channel x 512 Kbit (1692A)

A2 01680-66519 1 Acquisition Board 102 Channel x 512 Kbit (1691A)

A2 01680-66520 1 Acquisition Board 136 Channel x 512 Kbit (1690A)

A3 01680-66501 1 PCI Display Board

A4 01680-66504 1 Power Distribution Board

A5 01680-66530 1 CD-ROM Interface Board

A6 01680-83513 1 Hard Disk Drive with Software

A7 0950-2782 1 Flexible Disk Drive

A8 0950-3403 1 Power Supply

A9 0950-4068 1 Inverter

A10 0950-4373 1 CD-ROM Drive

A12 1150-7809 1 Mini Keyboard

A13 3160-0818 2 Fan

A14 0960-2453 1 PCI IEEE 1394 Board

A15 1150-7845 1 Mouse

E2 16542-61607 1 Double Probe Adapter

E3 5090-4833 8 Grabber Kit Assembly (1680A, AD)

Show 4 more labels / short text on this page
Replaceable Parts · Exchange Assemblies · 01680-69603 Motherboard Assembly · 16600-69600 Power Supply
Page 148 of the Agilent Technologies 1690 user manual — Agilent 1680A,AD-Series Replaceable Parts
Page 148 · click to enlarge
Page 148
Chapter 7Replaceable Parts

Ref. Des. Agilent Part Number QTY Description

E3 5090-4833 6 Grabber Kit Assembly (1681A, AD)

E3 5090-4833 4 Grabber Kit Assembly (1682A, AD)

E3 5090-4833 2 Grabber Kit Assembly (1683A, AD)

E4 5959-9333 0 Replacement Probe Leads (Qty 5)

E5 5959-9335 0 Replacement Pod Grounds - 5” (Qty 2)

E6 5959-9334 8 Probe Grounds - 2” (Qty 5, 1680A, AD)

E6 5959-9334 6 Probe Grounds - 2” (Qty 5, 1681A, AD)

E6 5959-9334 4 Probe Grounds - 2” (Qty 5, 1682A, AD)

E6 5959-9334 2 Probe Grounds - 2” (Qty 5, 1683A, AD)

H1 0380-1927 6 Hex Standoff 0.185 in (I/O panel to motherboard ports)

H2 0515-0365 3 M2.0 x 0.40; 4 mm T6 PH (CD-ROM drive to CD- ROM drive bracket)

H3 0515-0372 40 M3.0 x 0.50; 8 mm T10 PH (I/O panel to chassis, module cover to chassis, CD-ROM top bracket to CD-ROM bottom bracket, distribution board to chassis, probe shroud to acquisition board, acquisition board to chassis, probe shroud to chassis, PCI board and ISA slot covers to chassis, CD-ROM drive assembly to chassis, sleeve to chassis)

H4 0515-0433 12 M4.0 x 0.70; 8 mm T20 PH (power supply to chassis, cable tray to sleeve, rear feet to chassis)

H5 0515-1403 8 M4.0 x 0.70; 6 mm T15 90° FH (snap to sleeve (accessory pouch), front panel assembly to chassis)

H6 0515-1934 2 M2.5 x 0.45; 6 mm T8 PH (inverter to chassis)

H7 0515-1974 2 M2.5 x 0.45; 4 mm T8 PH (flexible disk drive to chassis)

H8 0515-2306 5 M3.0 x 0.5; 10 mm T10 PH (probe cable to probe shroud, 1680A, AD)

H8 0515-2306 4 M3.0 x 0.5; 10 mm T10 PH (probe cable to probe shroud, 1681A, AD)

H8 0515-2306 3 M3.0 x 0.5; 10 mm T10 PH (probe cable to probe shroud, 1682A, AD)

H8 0515-2306 2 M3.0 x 0.5; 10 mm T10 PH (probe cable to probe shroud, 1680A, AD)

H9 2360-0452 4 6-32 0.250 in T15 PH with washer (hard disk drive to chassis)

H10 2950-0054 2 Hex Nut 0.625 x 0.125; 1/2-28 (acquisition board BNC connectors to probe shroud)

Replaceable Parts

Page 149 of the Agilent Technologies 1690 user manual — Agilent 1680A,AD-Series Replaceable Parts
Page 149 · click to enlarge
Page 149
Chapter 7Replaceable Parts

Ref. Des. Agilent Part Number QTY Description

H11 0515-0430 6 M3.0 x 5.0; 6 mm T10; motherboard to chassis

MP1 01660-09101 4 Ground Spring (1680A, AD)

MP1 01660-09101 3 Ground Spring (1681A, AD)

MP1 01660-09101 2 Ground Spring (1682A, AD)

MP1 01660-09101 1 Ground Spring (1681A, AD)

MP2 01680-00201 1 I/O Panel

MP3 01680-01203 1 CD-ROM Bracket - Bottom

MP4 01680-01204 1 CD-ROM Bracket - Top

MP5 01680-04101 1 Rear Cover

MP6 01680-04701 1 Tilt Stand

MP7 01680-40401 1 Cable Tray

MP8 01680-41001 4 Rear Foot

MP9 01680-44101 1 Pod Cover (1681A, AD)

MP9 01680-44101 2 Pod Cover (1682A, AD)

MP9 01680-44101 3 Pod Cover (1683A, AD)

MP10 01680-52201 1 Front Cover

MP11 01680-60101 1 Chassis

MP12 01680-68701 1 Probe Shroud

MP13 01680-68702 1 Accessory Pouch

MP14 01680-68707 1 Handle Assembly

MP15 01680-68708 1 Sleeve Assembly

MP16 01680-87102 1 EMI Gasket (I/O panel to chassis)

MP17 01680-87103 EMI Gasket (probe shroud to chassis)

MP18 01680-88601 Adhesive stripe (EMI Gasket to probe shroud)

MP19 01680-94307 1 Cable Tray Label (1680A, AD)

MP19 01680-94308 1 Cable Tray Label (1681A, AD)

MP19 01680-94309 1 Cable Tray Label (1682A, AD)

MP19 01680-94310 1 Cable Tray Label (1683A, AD)

MP20 01680-94311 1 Label - Cable Installation

MP21 01680-94312 1 Label - Probe Shroud

MP22 01680-94313 1 Label - Pod and Cable

MP23 01680-94314 1 Label - Certification

MP24 01680-94315 1 Label - Rating

MP25 01680-94303 1 ID Label (1680A)

MP25 01680-94316 1 ID Label (1680AD)

MP25 01681-94301 1 ID Label (1681A)

Replaceable Parts

Page 150 of the Agilent Technologies 1690 user manual — Agilent 1680A,AD-Series Replaceable Parts
Page 150 · click to enlarge
Page 150
Chapter 7Replaceable Parts

Ref. Des. Agilent Part Number QTY Description

MP25 01681-94302 1 ID Label (1681AD)

MP25 01682-94301 1 ID Label (1682A)

MP25 01682-94302 1 ID Label (1682AD)

MP25 01683-94301 1 ID Label (1683A)

MP25 01683-94302 1 ID Label (1683AD)

MP26 0361-1787 8 Push Rivet (Fan to chassis)

MP27 0400-0727 0.35 Grommett, Stainless, Nylon Coated (front edge of power supply, chassis)

MP28 0400-0929 4 Snap (accessory pouch to sleeve)

MP29 0460-2010 Double-sided tape (flexible disk drive cable to flexible disk drive)

MP30 1400-0249 Cable Tie (motherboard cables, on/off cable)

MP31 1400-1254 2 Cable Clip 0.5 in diameter 0.75 in wide PVC (fan cables to chassis)

MP32 1400-2120 2 ISA Slot Cover

MP33 1400-3153 3 Cable Clamp (IEEE 1394 cable to CD-ROM assembly, IEEE 1394 cable to chassis)

MP34 54810-61001 4 Bottom Foot

MP35 8160-1545 0.65 EMI Gasket (I/O panel to chassis, CD-ROM assembly to chassis)

W1 01680-61604 1 PC Power Cable (distribution board to motherboard)

W2 01680-61605 1 Power Supply Cable (14- and 10- pin power supply output)

W3 01680-61606 1 Flexible Disk Drive Cable (distribution board to flexible disk drive)

W4 01680-61608 1 Inverter Cable (distribution board to inverter board)

W5 01680-61609 2 Fan Cable

W6 01680-61612 1 Flexible Disk Drive Cable (distribution board to motherboard)

W7 01680-61613 1 Disk Drive Data Cable (hard disk drive to motherboard)

W8 01680-61614 1 Hard Disk Drive Power Cable

W9 01680-61619 1 IEEE 1394 6-pin Cable (acquisition board to PCI IEEE 1394 board)

W10 01680-61622 1 Audio Cable (CD-ROM drive to motherboard)

W11 01680-61623 1 PC Sequence Cable (on/off, motherboard to distribution board)

Replaceable Parts

Page 151 of the Agilent Technologies 1690 user manual — Agilent 1680A,AD-Series Replaceable Parts
Page 151 · click to enlarge
Page 151
Chapter 7Replaceable Parts

Ref. Des. Agilent Part Number QTY Description

W12 16600-61602 1 Power Sense Cable (power supply to distribution board)

W13 16700-61604 1 Power Supply Cable (16-pin power supply output)

W14 16715-61601 4 Probe Cable (1680A, AD)

W14 16715-61601 3 Probe Cable (1681A, AD)

W14 16715-61601 2 Probe Cable (1682A, AD)

W14 16715-61601 1 Probe Cable (1683A, AD)

W15 54801-61611 1 Disk Drive Data Cable (CD-ROM drive to motherboard)

A16 01680-41901 1 Keypad*

A17 01680-66502 1 Front Panel Circuit Board

A18 01680-66506 1 Power Switch Interface Board

A19 2090-0827 1 LCD Display

H11 0515-0372 9 M.30 x 0.50; 8 mm T10 PH (front panel circuit board to front frame, LCD display to front frame)

H12 0515-1934 2 M2.5 x 0.45; 6 mm T8 PH (power switch interface board to front frame)

MP36 01680-01702 1 Module Cover

MP37 01680-41901 1 On/Off Keypad*

* The front panel Keypad and On/Off Keypad are ordered together as one unit. When the Keypads are received, detach and use what is needed.

MP38 01680-68712 1 Front Frame Assembly (includes the following

01680-94301 Front Panel Label

2190-0027 WIL.256 .478 .02 (ground lug to front frame)

2950-0072 NUTH 1/4-32 .062 (ground lug to front frame)

MP39 01680-68713 1 Lens Assembly (includes the following)

MP40 01680-94302 1 Module Cover Label

Show 7 more labels / short text on this page
Replaceable Parts · Front Panel Assembly · 01680-40501 Front Frame · 54542-46101 Ground Lug · 0360-1646 Terminal Jack · 01680-87101 Lens Gasket · 01680-88001 Lens
Page 152 of the Agilent Technologies 1690 user manual — Agilent 1680A,AD-Series Replaceable Parts
Page 152 · click to enlarge
Page 152
Chapter 7Replaceable Parts

Ref. Des. Agilent Part Number QTY Description

MP41 1400-0611 1 Cable Clamp, 1 in x 1 in adhesive backed (LCD to LCD display)

MP42 54801-47401 3 Knob 12 MM Flint Gray

MP43 54801-47402 3 Knob 18 MM Gray

W16 01680-61602 1 Display Cable

W17 01680-61603 1 On/Off Cable

W18 01680-61611 1 Calibration Cable

W19 01680-61615 1 Front Panel Cable

Replaceable Parts

Page 153 of the Agilent Technologies 1690 user manual — Exploded View
Page 153 · click to enlarge
Page 153
Chapter 7Replaceable Parts

Exploded View

Agilent Technologies 1690 — Exploded View diagram (page 153)
Fig. 112 · Exploded View

Exploded view of the Agilent 1690A,AD-series logic analyzer

Page 154 of the Agilent Technologies 1690 user manual — Agilent 1690A,AD-Series Replaceable Parts
Page 154 · click to enlarge
Page 154
Chapter 7Replaceable Parts

Agilent 1690A,AD-Series Replaceable Parts

Ref. Des. Agilent Part Number QTY Description

01680-69507 1 Acquisition Board 136 Channel x 2 Mbit (1690AD)

01680-69508 1 Acquisition Board 34 Channel x 2 Mbit (1693AD)

01680-69509 1 Acquisition Board 68 Channel x 2 Mbit (1692AD)

01680-69510 1 Acquisition Board 102 Channel x 2 Mbit (1691AD)

01680-69517 1 Acquisition Board 34 Channel x 512 Kbit (1693A)

01680-69518 1 Acquisition Board 68 Channel x 512 Kbit (1692A)

01680-69519 1 Acquisition Board 102 Channel x 512 Kbit (1691A)

01680-69520 1 Acquisition Board 136 Channel x 512 Kbit (1690A)

Replacement Parts A1 01680-66507 1 Acquisition Board 136 Channel x 2 Mbit (1690AD)

A1 01680-66508 1 Acquisition Board 34 Channel x 2 Mbit (1693AD)

A1 01680-66509 1 Acquisition Board 68 Channel x 2 Mbit (1692AD)

A1 01680-69510 1 Acquisition Board 102 Channel x 2 Mbit (1691AD)

A1 01680-69517 1 Acquisition Board 34 Channel x 512 Kbit (1693A)

A1 01680-69518 1 Acquisition Board 68 Channel x 512 Kbit (1692A)

A1 01680-69519 1 Acquisition Board 102 Channel x 512 Kbit (1691A)

A1 01680-69520 1 Acquisition Board 136 Channel x 512 Kbit (1690A)

A2 01680-66515 1 Distribution Board

A3 0950-4117 1 Power Supply (includes power supply output cables)

A4 3160-1006 2 Fan

E2 16542-61607 1 Double Probe Adapter

E3 5090-4833 8 Grabber Kit Assembly

E3 5090-4833 6 Grabber Kit Assembly

E3 5090-4833 4 Grabber Kit Assembly

E3 5090-4833 2 Grabber Kit Assembly

E4 5959-9333 0 Replacement Probe Leads (Qty 5)

E5 5959-9335 0 Replacement Pod Grounds - 5” (Qty 2)

E6 5959-9334 8 Probe Grounds - 2” (Qty 5)

E6 5959-9334 6 Probe Grounds - 2” (Qty 5)

E6 5959-9334 4 Probe Grounds - 2” (Qty 5)

E6 5959-9334 2 Probe Grounds - 2” (Qty 5)

Replaceable Parts · Exchange Assemblies

Page 155 of the Agilent Technologies 1690 user manual — Agilent 1690A,AD-Series Replaceable Parts
Page 155 · click to enlarge
Page 155
Chapter 7Replaceable Parts

Ref. Des. Agilent Part Number QTY Description

H1 0515-0372 31 M3.0 x 0.50; 8 mm T10 PH (distribution board to chassis, deck to chassis, probe shroud to acquisition board, acquisition board to deck, probe shroud to front panel, sleeve to chassis)

H2 0515-1035 2 M3.0 x 0.50; 8 mm T10 90° FH (line cable assembly to chassis rear)

H3 0515-1403 20 M4.0 x 0.70; 6 mm T15 90° FH (front panel to front frame, front frame to chassis, snap to sleeve (accessory pouch))

H4 0515-1934 2 M2.5 x 0.45; 6 mm T8 PH (power switch interface to fascia)

H5 0515-2306 5 M3.0 x 0.5; 10 mm T10 PH (probe cable to probe shroud, 1690A, AD)

H6 0515-2306 4 M3.0 x 0.5; 10 mm T10 PH (probe cable to probe shroud, 1691A, AD)

H6 0515-2306 3 M3.0 x 0.5; 10 mm T10 PH (probe cable to probe shroud, 1692A, AD)

H6 0515-2306 2 M3.0 x 0.5; 10 mm T10 PH (probe cable to probe shroud, 1693A, AD)

H7 54503-25701 2 Hex Nut (acquisition board BNC connectors to probe shroud)

MP1 01660-09101 4 Ground Spring (1690A, AD)

MP1 01660-09101 3 Ground Spring (1691A, AD)

MP1 01660-09101 2 Ground Spring (1692A, AD)

MP1 01660-09101 1 Ground Spring (1693A, AD)

MP2 01680-44101 1 Pod Cover (1691A, AD)

MP2 01680-44101 2 Pod Cover (1692A, AD)

MP2 01680-44101 3 Pod Cover (1693A, AD)

MP3 01680-68701 1 Probe Shroud

MP4 01680-68702 1 Accessory Pouch

MP5 01680-94313 1 Label - Pod and Cable

MP6 01680-94314 1 Label - Certification

MP7 01690-00101 1 Deck

MP8 01690-60101 1 Chassis

MP9 01690-60201 1 Front Panel

MP10 01690-68701 1 Sleeve Assembly

MP11 01690-68702 1 Fascia Assembly (includes the following)

01680-66506 Power Switch Interface Board

Replaceable Parts

Page 156 of the Agilent Technologies 1690 user manual — Agilent 1690A,AD-Series Replaceable Parts
Page 156 · click to enlarge
Page 156
Chapter 7Replaceable Parts

Ref. Des. Agilent Part Number QTY Description

0515-1934 M2.5 x 0.45; 6 mm T8 PH (power switch interface board to fascia)

MP12 01690-94302 1 Label - Probe Shroud

MP13 01690-94301 1 ID Label (1690A)

MP13 01690-94304 1 ID Label (1690AD)

MP13 01691-94301 1 ID Label (1691A)

MP13 01691-94302 1 ID Label (1691AD)

MP13 01692-94301 1 ID Label (1692A)

MP13 01692-94302 1 ID Label (1692AD)

MP13 01693-94301 1 ID Label (1693A)

MP13 01693-94302 1 ID Label (1693AD)

MP14 01690-94305 1 Warning Label

MP15 01690-94303 1 Rating Label

MP16 0361-1787 8 Push Rivet (Fan to Chassis)

MP17 0400-0929 4 Snap (accessory pouch to sleeve)

MP18 1400-1254 3 Cable Clip 0.5 in diameter 0.75 in wide PVC (fan cables to chassis, on/off cable to chassis)

MP19 16600-49301 2 Molded Plastic Bumper

MP20 16600-68707 1 Handle Assembly

MP21 5022-1188 1 Front Frame

MP22 5041-9171 2 Side Trim

MP23 5041-9176 1 Top Trim

MP24 54503-25701 2 Hex Nut

MP25 54810-61001 4 Bottom Foot

W1 01680-61609 2 Fan Cable

W2 01690-61601 1 Line Cable Assembly (chassis rear to power supply)

W3 01690-61603 1 Power Sense Cable (power supply to distribution board)

W4 01690-61604 1 Power Supply On/Off Cable (power supply to distribution board)

W5 16715-61601 4 Probe Cable (1690A, AD)

W5 16715-61601 3 Probe Cable (1691A, AD)

W5 16715-61601 2 Probe Cable (1692A, AD)

W5 16715-61601 1 Probe Cable (1693A, AD)

Show 4 more labels / short text on this page
Replaceable Parts · 01690-41901 On/Off Keypad · 01690-44101 Fascia · 01690-61602 On/Off Cable
Page 157 of the Agilent Technologies 1690 user manual — Power Cables and Plug Configurations
Page 157 · click to enlarge
Page 157
Chapter 7Replaceable Parts

Power Cables and Plug Configurations

This instrument is equipped with a three-wire power cable. The type of power cable plug shipped with the instrument depends on the country of destination.

Plug Type Cable Part No.

Plug Description Length (in/cm)

Opt 900 250V 8120-1703 90° 90/228 Mint Gray United Kingdom, Cyprus, Nigeria, Zimbabwe, Singapore

Agilent Technologies 1690 — Power Cables and Plug Configurations diagram (page 157)
Fig. 113 · Power Cables and Plug Configurations

Opt 901 250V 8120-0696 90° 87/221 Mint Gray Australia, New Zealand

8120-1692 90° 79/200 Mint Gray East and West Europe, Saudi Arabia, So. Africa, India (unpolarized in many nations)

8120-1521 90° 90/228 Jade Gray United States, Canada, Mexico, Philippines, Taiwan

8120-6799 90° 90/228 Israel

Opt 920 250 V

8120-2296 1959-24507 Type 12 90°

79/200 Mint Gray Switzerland

Opt 912 250V 8120-2957 90° 79/200 Mint Gray Denmark

Opt 917 250V 8120-4600 90° 79/200 Republic of South Africa India

Show 6 more labels / short text on this page
Color Country · Opt 902 250V · Opt 903** 125V · Opt 919 250V · 8120-6871 90° Argentina · Opt 906 250V
Page 158 of the Agilent Technologies 1690 user manual — Power Cables and Plug Configurations
Page 158 · click to enlarge
Page 158
Chapter 7Replaceable Parts

Plug Type Cable Part No.

Plug Description Length (in/cm)

8120-4754 90° 90/230 Japan

Opt 921 8120-6979 90° Chile

Opt 922 8120-8377 90° People’s Republic of China

Opt 927 8120-8871 90° Thailand

* Part number shown for plug is industry identifier for plug only. Number shown for cable is Agilent Technologies part number for complete cable including plug. ** These cords are included in the CSA certification approval of the equipment. E = Earth Ground L = Line N = Neutral

Color Country · Opt 918 100V

Page 159 of the Agilent Technologies 1690 user manual — Theory of Operation
Page 159 · click to enlarge
Page 159

Theory of Operation

This chapter tells the theory of operation for the logic analyzer and describes the self-tests.

Page 160 of the Agilent Technologies 1690 user manual — Block-Level Theory
Page 160 · click to enlarge
Page 160
Chapter 8Theory of Operation

The information in this chapter will help you understand how the logic analyzer operates and what the self-tests are testing. This information is not intended for component-level repair.

Block-Level Theory

The block-level theory is divided into two parts: theory for the logic analyzer and theory for the acquisition boards. A block diagram is shown with each theory.

Agilent Technologies 1690 — Block-Level Theory diagram (page 160)
Fig. 114 · Block-Level Theory

Agilent 1680A,AD-Series Logic Analyzer Block Diagram

Page 161 of the Agilent Technologies 1690 user manual — Agilent 1680A,AD-series Logic Analyzer Theory
Page 161 · click to enlarge
Page 161
Chapter 8Theory of Operation Agilent 1680A,AD-series Logic Analyzer Theory

Agilent 1680A,AD-series Logic Analyzer Theory

PC Motherboard. The Agilent 1680A,AD-series benchtop analyzer is built around an x86 ATX motherboard. The motherboard serves as the system backplane through PCI slots and IEEE 1394 ports. The hard drive, flexible disk drive, and communications ports are all integrated into the 1680A,AD-series logic analyzer through the PC motherboard. PCI slots are used to house LAN, video, and the IEEE 1394 interface.

PC Software System. The user interface and I/O run on the PC motherboard under Microsoft Windows ® 2000 Professional operating system. This is the primary software system. Windows ® 2000 Professional operating system. This is the primary software system. Windows ® 2000 Professional provides all the graphics drivers needed to create the front panel interface as well as providing all of the software necessary to access the disk drives, to drive printers, and to process input from pointing devices and the keyboard. Standard Windows ® drivers are also used for I/O boards installed in the PCI slots and devices connected to LAN, video, and the IEEE 1394 interface.

Disk Drives. The Agilent 1680A,AD-series logic analyzer contains a 3.5” hard disk drive as well as a 3.5” micro flexible disk drive.

Flat Panel Display. The Agilent 1680A,AD-series logic analyzer includes an active matrix thin flim transistor (AM-TFT) liquid crystal color flat panel display (LCD). The LCD had a resolution of 800 x 600 (SVGA resolution) that measures 12.1 diagonally.

Luminance is software selectable at either 60% luminance or 100% luminance. The display’s luminance is controlled from the user interface by sending commands to the IEEE 1394 processor which in turn sends commands to the front panel microcontroller where luminance is controlled. This board plugs into a PCI slot on the PC motherboard and has a cable that drives the display.

An inverter is used to provide power to the two fluorescent lamps that make up the backlight of the liquid crystal display. The inverter’s output voltage is derived from a +12 V input. The inverter is a separate OEM component that is powered from the power distribution board.

A custom SVGA display board plugs into a PCI slot. The SVGA board drives the flat panel display and can also drive a standard external CRT display.

Networking. Networking is accomplished using an OEM board that plugs into a PCI slot.

Other I/O. The parallel port, audio port, keyboard port, mouse port, serial port, and USB ports are all integrated onto the motherboard.

Page 162 of the Agilent Technologies 1690 user manual — Acquisition Board
Page 162 · click to enlarge
Page 162
Chapter 8Theory of Operation Agilent 1680A,AD-series Logic Analyzer Theory

Power Supply

A low voltage power supply provides all the DC voltages needed to operate the logic analyzer. The power supply also provides the +5V VDC voltage to the probe cables to power the logic analyzer accessories and analysis probes.

Unfiltered voltages of +12V, -12V, +5V, -5.2V, +3.4 V and +5V (standby) are supplied to the power distribution board where they are filtered and distributed to other boards and components in the system.

Acquisition Board

Logic Acquisition Board Block Diagram

Agilent Technologies 1690 — Acquisition Board diagram (page 162)
Fig. 115 · Acquisition Board

Probing. The probing system consists of a tip network, a probe cable, and terminations that reside on the analyzer board. Each probe cable is made up of two ribbonized coaxial cables carrying 16 data channels and 1 clock/data channel.

Each channel of the probing system had its own ground. In addition, the pod has a single ground. For applications where many channels are used (greater than three) and signal times are less than 3 ns, individual channel grounds should be used.

Page 163 of the Agilent Technologies 1690 user manual — Comparators
Page 163 · click to enlarge
Page 163
Chapter 8Theory of Operation Agilent 1680A,AD-series Logic Analyzer Theory

The probe tip networks are comprised of a series of resistors (250 Ohm) connected to a parallel combination of a 90 KOhm resistor and an 8.5 pF capacitor. The parallel 90 KOhm and 8.5 pF capacitor along with the glossy cable and terminations form a divide-by Ohm tip resistor is used to buffer (or raise the impendence of) the 8.5 pF capacitor that is in series with the cable capacitance.

Comparators. Two 9-channel comparators interpret the incoming data and clock signals as either high or low, depending on where the user-programmable threshold is set. The threshold voltage of each pod is individually programmed, and the voltage selected applies to the clock channel as well as the data channels of each pod.

Each of the comparators has a serial test input port used for testing purposes. A test bit pattern is sent from the Test and Clock Synchronization Circuit to the comparator. The comparators then propagate the test signal on each of the nine channels of the comparator. Consequently, the operating system software can test all data and clock channel pipelines on the circuit board through the comparator.

Acquisition. Each acquisition circuit is made up of a single acquisition circuit. Each acquisition is a 34-channel state/timing analyzer. One to four acquisition ICs are included on each logic analyzer board for a total of up to 128 data channels and four state clock (pods one through four) in state mode. There are 136 data channels available in timing mode. All of the sequencing, store qualification, pattern/range recognition and event counting functions are performed by the acquisition IC.

Additionally, the acquisition ICs perform master clocking functions. All four state acquisition clocks are sent to the first two acquisition ICs, and the acquisition ICs generate their own sample clocks. When necessary, the acquisition ICs individually perform a clock optimization after the user selects the RUN icon and before data is stored.

Clock optimization involves using programmable delays in the acquisition ICs to position the master clock transition where valid data is captured. This procedure greatly reduces the effects of channel-to-channel skew and other propagation delays.

In the timing acquisition mode, an oscillator-driven clock circuit provides a 100 MHz clock signal to each of the acquisition IPs where they are multiplied by a PLL to obtain the necessary internal clock frequency. For high speed timing acquisition, the master acquisition IC derives the sampling frequency using its PLL and redistributes this sampling clock to the other acquisition ICs.

Acquisition RAM. The acquisition RAM is external to the acquisition IC. The acquisition RAM consists of 9 RAM ICs per acquisition chip. A CPLD, which is initialized by the FPGA, increments the memory addresses while reading or writing to the memory. Memory is read to the FPGA where it is translated and

Page 164 of the Agilent Technologies 1690 user manual — Test and Clock Synchronization Circuit
Page 164 · click to enlarge
Page 164
Chapter 8Theory of Operation Agilent 1680A,AD-series Logic Analyzer Theory

resent via DMA transactions to an IEEE 1394 Link Layer chip. The Link Layer then transmits the data to a 1394 PHY (physical layer chip) where the data is transmitted over a 1394 cable to the motherboard for processing.

Test and Clock Synchronization Circuit. ECLinPS (ECL in pico seconds) ICs are used in the Test and Clock Synchronization Circuit for reliability and low channel-to-channel skew. Test patterns are generated and sent to the comparators during software operation verification (self-tests). The test patterns are propagated across all data and clock channels and read by the acquisition ICs to verify that the data and clock pipelines are operating correctly.

Clock and Data Threshold. The threshold circuit includes a precision octal DAC. Each of the eight channels of the DAC is individually programmable which allows the user to set the thresholds of the individual pods. The 16 data channels and the clock/data channel of each pod are all set to the same threshold voltage.

Motherboard Interface. The motherboard communications to the acquisition board over an IEEE 1394 interface residing on the acquisition board. Changes to the logic analyzer configuration made in application software are translated into configuration commands and then sent to the acquisition board through this interface. All state and timing functions including storage qualification, sequencing, assigning clocks and qualifiers, RUN and STOP, and thresholds are controlled in the manner.

A microcontroller manages initialization of the acquisition board at power-up, reconfiguring the acquisition board as a result of user input, and managing the IEEE 1394 communication to and from the motherboard.

A field programmable gate array (FPGA) bridges the 1394 interface to the rest of the acquisition board components. It also serves as the memory controller for the acquisition memory.

Memory Address Counters (MACs). Each acquisition IC has a CPLD that is used to provide addresses to the memory ICs that are written during an acquisition. The MACs are also used when uploading data to the GUI. Each CPLD contains three MACs. The MACs are configured serially by the FPGA prior to each acquisition and prior to each data upload. The application is responsible for setting up the proper address by writing to various register in the FPGA, which results in the MACs being serially programmed by the FPGA.

+5 VDC supply. The +5 VDC supply circuit supplies power to active logic analyzer accessories such as analysis probes. Thermistors on the +5 VDC supply lines protect the logic analyzer and the active accessory from overcurrent conditions. When an overcurrent condition is sensed, the thermistors create an open that shuts off the current rom the +5 VDC supply. After the overcurrent condition is resolved the thermistor closes the circuit and makes the supply current available.

Page 165 of the Agilent Technologies 1690 user manual — Front Panel Board
Page 165 · click to enlarge
Page 165
Chapter 8Theory of Operation Agilent 1680A,AD-series Logic Analyzer Theory

Power Distribution Board

The power distribution board connects directly to the power supply and distributes power to the rest of the boards in the system including the motherboard. It also distributes power to the disk drives, fan, and CD-ROM. It has circuity for regulating fan voltage that is temperature dependent as well as detecting when a fan ceases to spin.

The board also converts standard ATX IDE interface signals into those used by notebook flexible disk drives and hard disk drives if needed in the system. The power distribution also distributes various signals between boards such as serial lines and front panel ID signals that connect between the acquisition board and the front panel board.

Front Panel Board

The front panel board contains a microcontroller that is powered from 5V. This particular 5V line comes from a standby rail that is supplied as long as the unit is connected to AC power. The standby rail powers the microcontoller and some core circuitry on the front panel board. The remaining circuitry on the front panel board is powered by the main 5V rail after the supply is turned on by the front panel microcontroller.

The standby power also supplies a small portion of the motherboard circuitry. When the motherboard detects that the power switch has been pressed, it will assert a signal to the microcontroller (via the power distribution board) that tells the front panel microcontroller to turn the main rails of the power supply on or off. After the system’s main rails have been powered up, the front panel microcontroller is then free to turn on front panel LED’s scan the keypad, and drive control signals to the inverter that cause the display to dim or turn on/off.

The logic analyzer application communicates with the front panel by sending commands to the acquisition microcontroller, via the 1394 bus, which are then transmitted serially by the acquisition microcontroller to the front panel microcontroller. Data is transmitted in the reverse order when the application is polling the front panel.

Page 166 of the Agilent Technologies 1690 user manual — Power Distribution Board
Page 166 · click to enlarge
Page 166
Chapter 8Theory of Operation Agilent 1690A,AD-series Logic Analyzer Theory

Agilent 1690A,AD-series Logic Analyzer Theory

Acquisition Board

The Agilent 1690A,AD-series logic analyzers utilize the same acquisition board as the 1680A,AD-series benchtop logic analyzers. The motherboard interface connects directly to the IEEE 1394 port on the host PC.

Power Supply

A low voltage power supply provides all the DC voltages needed to operate the logic analyzer acquisition board. The power supply also provides the +5V VDC voltage to the probe cables to power the logic analyzer accessories and analysis probes.

Unfiltered voltages of +12V, +5V, -5.2V, and +3.4 are supplied to the power distribution board where they are filtered and distributed to other boards and components in the system.

Power Distribution Board

The power distribution board connects directly to the power supply and distributes power to the rest of the system. The power distribution board has circuitry for regulating fan voltage that is temperature dependent as well as detecting when a fan ceases to spin.

Page 167 of the Agilent Technologies 1690 user manual — Connectivity Tests (1690A,AD-series)
Page 167 · click to enlarge
Page 167
Chapter 8Theory of Operation

Self-Tests Descriptions

The self-tests identify the correct operation of major functional areas in the logic analyzer. The self-tests are not intended for component-level diagnostics.

Three types of tests are performed on the Agilent 1680A,AD- and 1690A,AD- series logic analyzers: the power-up self-tests, the functional performance verification self-tests, and the parametric performance verification tests.

The power-up self-tests are performed when power is applied to the instrument.

The functional performance verification self-tests are run using a separate operating system, the performance verification (PV) operating system.

Parametric performance verification requires the use of external test equipment that generates and monitors test data for the logic analyzer to read. The performance verification procedures in chapter 3 of this service guide make up the parametric performance verification for the logic analyzer. Refer to chapter 3, "Testing Performance," for further information about parametric performance verification.

Power-up Self-Tests (1680A,AD-series)

The power-up self tests on the 1680A,AD-series logic analyzers is performed by the Microsoft Windows ® 2000 Professional operating system. As part of the Windows 2000 Professional power on self test (POST), the presence of all required system components is verified. When the text "Starting Windows . . ." appears, this means required system components have been detected and have passed their individual power-up self-tests.

For more information on the Windows ® 2000 Professional POST, refer to Microsoft on-line(http://www.microsoft.com) and the Microsoft Support Services Knowledge Base (http://search.support.microsoft.com/kb/).

Connectivity Tests (1690A,AD-series)

A Connectivity test routine is done on an Agilent 1690A,AD-series hosted logic analyzer when it is connected to a host PC. A communications test is first done on the hosted logic analyzer when the IEEE 1394 plug and play interface is connected between the host PC and the logic analyzer. After the communications test is successfully completed, an operational accuracy calibration is done on the logic analyzer to test the major subsystems of logic analyzer acquisition and to initialize the acquisition system.

Self-Tests Descriptions

Page 168 of the Agilent Technologies 1690 user manual — Acquisition Board Self Tests
Page 168 · click to enlarge
Page 168
Chapter 8Theory of Operation Self-Tests Descriptions

Acquisition Board Self Tests

The acquisition board self tests are available in the Agilent Logic Analyzer application software user interface. These self tests verify the correct operation of the acquisition board in both the Agilent 1680A,AD-series and 1690A,AD- series logic analyzers

Register Test. The Register Test verifies that the registers of each acquisition IC are operating properly. Test patterns are written to each register on each acquisition IC, read, and compared with known values. The registers are reset, and verified that each register has been initialized. Test patterns are then written to ensure the chip address lines are not shorted or opened. Finally test data is written to registers of individual acquisition ICs to ensure each acquisition IC can be selected independently.

Passing the Register Test implies that the acquisition IC registers can store acquisition control data to properly manage the operating of each IC.

Memory Test. The Memory Test verifies that each bit in the acquisition memory IC can be written with a logic “0” and logic “1” through the Serial Access Memory port. Test data is generated using a shifting test register in the acquisition ICs. The serialized test patterns are then sent to the memory port of each acquisition memory IC and stored. The data in the acquisition memory ICs are then downloaded and compared with known values.

Passing the Memory Test implies the acquisition memory can store data written through the memory port. This test along with the Memory Modes Test provides complete testing of the memory ICs.

Comparator Test. The Comparator Test ensures the data signal comparators in the module front end can be set to their maximum and minimum thresholds and that they recognize activity at the signal inputs. A clock signal is routed to a test port on each comparator. The threshold is then set to the minimum value. The comparator output is then read and compared with a known value. The threshold is then set to a maximum value. The comparator output is again read and compared with a known value.

Passing the Comparators Test implies that the front-end comparators are operating properly, can recognize both a logic ì0î and logic ì1î, and can properly send the acquisition data downstream to the acquisition ICs.

Trigger Bus Test. The Trigger Bus Test verifies the trigger resource lines that run between each acquisition IC. The test ensures that the trigger resource lines can be both driven as outputs and read as inputs. The resource registers are written with test patterns, read back, then compared with known values. The resource registers are then written with test patterns, read back from a different acquisition IC, then compared with known values.

Page 169 of the Agilent Technologies 1690 user manual — Logic Analyzer Self-Tests
Page 169 · click to enlarge
Page 169
Chapter 8Theory of Operation

Trigger Arm Test. The Trigger Arm Test verifies that the local arm signal can be received by the master acquisition IC on the acquisition board. The test also verifies the global arm signal can be driven by each acquisition IC on a master board and received by all acquisition ICs on the card. The arm lines are asserted and read at the acquisition ICs to ensure each acquisition IC recognizes the signal.

Passing the Trigger Arm Test implies any acquisition IC can arm the card and that all acquisition ICs can recognize the arm signal.

Clock Paths Test. The Clock Paths Test verifies that the system Master, Slave, and Psync clocks are functional between the acquisition ICs. The module is configured to take a simple measurement. Test data is then created at the comparators and an acquisition taken. The resulting data is then downloaded and compared with known values.

Passing the Clock Paths Test implies that all acquisition IC clock lines can be driven by each acquisition IC and can be received by each acquisition IC in the module. Consequently each acquisition IC can reliably acquire data in response to the acquisition clock signal.

Memory Modes Test. The Memory Modes Test verifies the CPU interface can properly manage the acquisition memory unload in full channel, half-channel, count only, and interleaved modes. Test data is written to acquisition memory. Different unload modes are selected, then the data is read and compared with known values.

Passing the Memory Modes Test implies that the data can be reliably read from acquisition memory in full channel, half-channel, count only, and interleaved mode. This test along with the Memory Test provides complete testing of acquisition memory downloading through the 1394 interface.

Calibration Test. The Calibration Test ensures that each acquisition IC in the module can perform an operational accuracy self-calibration. Various self- calibration routines are initiated. The results of each self-calibration routine are then checked to see if the self-calibration was successful or not.

Passing the Calibration Test implies that the module can reliably perform an operation accuracy self-calibration. Consequently the incoming data path is optimized to reduce channel-to-channel skew so the acquisition ICs can reliably capture the incoming data.

Logic Analyzer Self-Tests

Register Test. The Register Test verifies that the registers of each acquisition IC is operating properly. Test patterns are written to each register on each acquisition IC, read, and compared with know values. The registers are reset, and

Self-Tests Descriptions

Page 170 of the Agilent Technologies 1690 user manual — Memory Test
Page 170 · click to enlarge
Page 170
Chapter 8Theory of Operation Self-Tests Descriptions

verified that each register has been initialized. Test patterns are then written to ensure the chip address lines are not shorted or opened. Finally test data is written to registers of individual acquisition ICs to ensure each acquisition IC can be selected independently.

Passing the Register Test implies that the acquisition IC registers can store acquisition control data to properly manage the operating of each IC.

Memory Test. The Memory Test verifies that each bit in the acquisition memory IC can be written with a logic “0” and logic “1” through the Serial Access Memory port. Test data is generated using a shifting test register in the acquisition ICs. the serialized test patterns are then sent to the memory port of each acquisition memory IC and stored. The data in the acquisition memory ICs are then downloaded and compared with known values.

Passing the Memory Test implies the acquisition memory can store data written through the memory port. This test along with the Memory Modes Test provides complete testing of the memory ICs.

Comparator Test. The Comparator Test ensures the data signal comparators in the module front end can be set to their maximum and minimum thresholds and that they recognize activity at the signal inputs. A clock signal is routed to a test port on each comparator. The threshold is then set to the minimum value. The comparator output is then read and compared with a known value. The threshold is then set to a maximum value. The comparator output is again read and compared with a known value.

Passing the Comparators Test implies that the front-end comparators are operating properly, can recognize both a logic “0” and logic “1”, and can properly send the acquisition data downstream to the acquisition ICs.

Trigger Bus Test. The Trigger Bus Test verifies the trigger resource lines that run between each acquisition IC. The test ensures that the trigger resource lines can be both driven as outputs and read as inputs. The resource registers are written with test patterns, read back from a different acquisition IC, then compared with known values.

Trigger Arm Test. The Trigger Arm Test verifies that the local arm signal can be received by the master acquisition IC on the acquisition board. The test also verifies the global arm signal can be driven by each acquisition IC on a master board and received by all acquisition ICs on the card. The arm lines are asserted and read at the acquisition ICs to ensure each acquisition IC recognized the signal.

Passing the Trigger Arm Test implies any acquisition IC can arm the card and that all acquisition ICs can recognize the arm signal.

Clock Paths Test. The Clock Paths Tests verifies that the system Master, Slave, and Psyn clocks are functional between the acquisitions ICs. The module us

Page 171 of the Agilent Technologies 1690 user manual — Memory Modes Test
Page 171 · click to enlarge
Page 171
Chapter 8Theory of Operation

configured to take a simple measurement. Test data is then created at the comparators and an acquisition taken. The resulting data is then downloaded and compared with known values.

Passing the Clock Path Test implies that all acquisition IC clock lines can be driven by each acquisition IC and can be received by each acquisition IC in the module. Consequently each acquisition IC can reliably acquire data in response to the acquisition clock signal.

Memory Modes Test. The Memory Modes Test verifies the CPU interface can properly manage the acquisition memory unload in full-channel, half-channel, count only, and interleaved modes. Test data is written to acquisition memory. Different unload modes are selected, then the data is read and compared with known values.

Passing the Memory Modes Test implies that the data can be reliably read from acquisition memory in full-channel, half-channel, count only, and interleaved mode. This test along with the Memory Test provides complete testing of acquisition memory downloading through the 1394 interface.

Calibration Test. The Calibration Test ensures that each acquisition IC in the module can perform an operational accuracy self-calibration. Various self- calibration routines are initiated. The results of each self-calibration routine are then checked to see if the self-calibration was successful or not.

Passing the Calibration Test implies that the module can reliably perform an operational accuracy self-calibration. Consequently the incoming data path is optimized to reduce channel-to-channel skew so the acquisition ICs can reliably capture the incoming data.

Self-Tests Descriptions

Page 172 of the Agilent Technologies 1690 user manual — Memory Modes Test
Page 172 · click to enlarge
Page 172
Chapter 8Theory of Operation Self-Tests Descriptions
Page 173 of the Agilent Technologies 1690 user manual — Warranty
Page 173 · click to enlarge
Page 173

Notices

© Agilent Technologies, Inc. 2001, 2004-2005

Manual Part Number

Technology Licenses

Safety Notices

The hardware and/or soft- ware described in this docu- ment are furnished under a license and may be used or copied only in accordance with the terms of such license.

A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.

No part of this manual may be reproduced in any form or by any means (including elec- tronic storage and retrieval or translation into a foreign lan- guage) without prior agree- ment and written consent from Agilent Technologies, Inc. as governed by United States and international copy- right laws.

Edition

01680-97003, October 2001 01680-97010, June 2004 01680-97011, August 2004 01680-97014, March 2005 01680-97016, October 2005

Restricted Rights Legend

Agilent Technologies, Inc. 1900 Garden of the Gods Road Colorado Springs, CO 80907 USA

If software is for use in the performance of a U.S. Govern- ment prime contract or sub- contract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June 1995), or as a “commer- cial item” as defined in FAR 2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is sub- ject to Agilent Technologies’ standard commercial license terms, and non-DOD Depart- ments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Lim- ited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applica- ble in any technical data.

Trademark Acknowledgements

Windows and MS Windows are U.S. registered trademarks of Microsoft Corporation.

Warranty

The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by appli- cable law, Agilent dis- claims all warranties, either express or implied, with regard to this manual and any information con- tained herein, including but not limited to the implied warranties of merchantability and fit- ness for a particular pur- pose. Agilent shall not be liable for errors or for inci- dental or consequential damages in connection with the furnishing, use, or performance of this document or of any infor- mation contained herein. Should Agilent and the user have a separate writ- ten agreement with war- ranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agree- ment shall control.

Windows XP is a U.S. registered trademark of Microsoft Corpora- tion.

A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.

Agilent Technologies P.O. Box 2197 1900 Garden of the Gods Road Colorado Springs, CO 80901

01680-97016 · Printed in Malaysia

Page 174 of the Agilent Technologies 1690 user manual — Memory Modes Test
Page 174 · click to enlarge
Page 174
Page 174 of the Agilent Technologies 1690 user manual — Memory Modes Test
Sources

References & links

Pages referenced in this manual — manufacturer, support and downloads.

Ask about the Agilent Technologies 1690

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Agilent Technologies 1690

Be the first to ask — other owners or our team will answer.

Original document

Agilent Technologies 1690 manual PDF

174 pages·11.1 MB·Weather Radio·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…