Instruction/ maintenance manual of the product THCD-100 Teledyne
Go to page of 28
INSTRUCTION MANUAL THCD-100 CONFIGURABLE DISPLAY TELEDYNE HASTINGS INSTRUMENTS ISO 9001 CER TIFIED.
165-072011 TH CD-100 Power Suppl y Page 2 of 28 Manual Print History The print history shown below lists the printing dates of all revisions and a ddenda created for this manual . The revision level increases alphabetically as the ma nual undergoes subseque nt updates.
165-072011 TH CD-100 Power Suppl y Page 3 of 28 Table of Contents 1. GENERAL INFO RMATIO N ............................................................................................................................................ 4 1.1. F EATURES ..
165-072011 TH CD-100 Power Suppl y Page 4 of 28 1. GENERAL INFORMA TION The THCD-100 is a multi featured process display controller, capable of interf acing dir ectly to a mass flow meter, mass flow controller or another proce ss transducer such as the THPS-760S Vacuum Transducer.
165-072011 TH CD-100 Power Suppl y Page 5 of 28 2. Inst allation and Interconnections 2.1. Installation The THCD-100 can be panel mounted, or used on a bench top and is a standard 96 x 48mm panel meter enclosure. Note that the in strument must always be situated in such a wa y as to enable adequate air circulation about the unit.
165-072011 TH CD-100 Power Suppl y Page 6 of 28 2.3. Interface Socket Pin out – Miniature 15 Pin 'D' Type 2.4. Serial Interface Socket Pinout – 9 Pin 'D' Type 1 For RS232 only, connect only the pins designated in bold font . 2 For RS485 only, connect only the pins designated in normal font.
165-072011 TH CD-100 Power Suppl y Page 7 of 28 2.5. Power Connection Socket The THCD-100 requires a 24V, 500mA supply to operate. This must be a rectified and regulated di rect current supply, capable of 15W (recommended).
165-072011 TH CD-100 Power Suppl y Page 8 of 28 3. FEA TURES 3.1. Introduction This section describes the principles used in the THCD-100. It de tails the various settings for the inputs and the outputs including setpoint control. It does no t detail any servicing or calibratio n procedures.
165-072011 TH CD-100 Power Suppl y Page 9 of 28 (assuming setpoint is in AU TO mode and not a slave source), then the output voltage of the setpoint woul d be 0.
165-072011 TH CD-100 Power Suppl y Page 10 of 28 b) Filter Size The size of the filter is currently configurable between 0 an d 6 seconds and simply indicates the amount of readings that are taken and b uffered to calculate and show the avera ged reading.
165-072011 TH CD-100 Power Suppl y Page 11 of 28 4. Front Panel Operation 4.1. Introduction The front panel consists of a 112x16 dot matrix V FD di splay with 6 membrane key switches below it.
165-072011 TH CD-100 Power Suppl y Page 12 of 28 4.3. Front Panel Screens Here follows a complete breakd own of all THCD-100 screens: a) Main Screen (not ch anging setpoint) The main screen comes in two forms. The fi rst shows the main channel input data (with range, fullscale and calibration applied) and the associated units string.
165-072011 TH CD-100 Power Suppl y Page 13 of 28 c) Main Menu Screen The Main Menu screen provides a path to select sc reens that are used for the configuration of the THCD-100 – the setpoint (value.
165-072011 TH CD-100 Power Suppl y Page 14 of 28 Input Screen This screen allows you to change the input c han nel range and full scale voltage (see the Principles section (5.2) for more informatio n on this) and also assign a un its label (up to 5, non-scaling characters) for the channel data.
165-072011 TH CD-100 Power Suppl y Page 15 of 28 f) Comms Menu Screen The Comms Menu screen provides opti ons for setting the serial communication’s bau d rate, address and RS232/RS485 protocol.
165-072011 TH CD-100 Power Suppl y Page 16 of 28 h) Relay Screen This screen allows you to set the trip point for when the relay switches. The relay also contains some hysteresis to st op it ‘chattering’. This i s configurable from this screen as well and is represented as a percentage of full scale of th e input channel, with limit s of 0.
165-072011 TH CD-100 Power Suppl y Page 17 of 28 5. SERIAL COMMUNICA TION 5.1. Introduction The THCD-100 allows full control and feedback via se rial communication protocols (RS232 or RS485). Calibration, in fact, all controls available through the fr ont panel, are available as commands sent using serial communications.
165-072011 TH CD-100 Power Suppl y Page 18 of 28 … where: <a> is the address letter <cmd> is the command/query (query identi fied by a ‘?’ after the three chars) <params> is the comm a separated parameter list, if any <response> is the accept ance indicator.
165-072011 TH CD-100 Power Suppl y Page 19 of 28 the data will be taken every 100ms and buffered up , to be sent out every 500 ms in blocks of 5 readings. This has proven to be a more robust output method. c. Setpoint Value – ‘spv’ This command is used to set the value of the command setpoint.
165-072011 TH CD-100 Power Suppl y Page 20 of 28 g. Setpoint Initial Mode – ‘sim’ As with the value, the setpoint mode set using the ‘spm’ command (above) is a volatile setting. Again this command provides a way of setting an initial st artup mode for the setpoint on power up of the THCD.
165-072011 TH CD-100 Power Suppl y Page 21 of 28 the new address. All subsequen t communications must use the new address. The current communication address can be obtained usin g the query ‘add? ’. The returned da ta is: ADDR: <address> 5.5.
165-072011 TH CD-100 Power Suppl y Page 22 of 28 situation and it therefore makes no sense to atte mpt to change the ba nd setting. For more information on this see the Principles section (5.6). To retrieve the current filter band, use the query ‘flb?’.
165-072011 TH CD-100 Power Suppl y Page 23 of 28 Other Commands a) User Input Re-zero – ‘irz’ To perform a user input re-zero func tion, use th is command. The co mmand needs no parameters. To find the current user re-zero values f or all input channels, use the query ‘irz? ’.
165-072011 TH CD-100 Power Suppl y Page 24 of 28 6. SER VICE & CALIBRA TION 6.1. Service There are no user serviceable parts in side the instruments. Should any di fficulties be encountered in the use of the THCD-100, it is recommended that you contact Hastings Instruments for advice and instruc tions.
165-072011 TH CD-100 Power Suppl y Page 25 of 28 7. Appendices 7.1. Appendix A Table 1 – THCD-100 Command Set Summary Purpose Cmd Query Parameters Notes Output current readings r Output current read.
165-072011 TH CD-100 Power Suppl y Page 26 of 28 7.2. Appendix B Table 2 – Factory Defaults Setting Default Value Additional Information Repeat Rate (for output readings) 0 Repeat off Serial Communi.
165-072011 TH CD-100 Power Suppl y Page 27 of 28 7.3. Appendix C Table 3 – Settings in ‘ras’ command string Description Type and size Input channel units string 5 char string range 8 char string.
165-072011 TH CD-100 Power Suppl y Page 28 of 28 8. W ARRANTY 8.1. Warranty Repair Policy H a s t i n g s I n s t r u m e n t s w a r r a n t s t h i s p r o d u c t f o r a p e r i o d o f o n e y e a r f r o m t h e d a t e o f s h i p m e n t t o b e f r e e from defects in material and workmanship.
An important point after buying a device Teledyne THCD-100 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Teledyne THCD-100 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Teledyne THCD-100 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Teledyne THCD-100 you will learn all the available features of the product, as well as information on its operation. The information that you get Teledyne THCD-100 will certainly help you make a decision on the purchase.
If you already are a holder of Teledyne THCD-100, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Teledyne THCD-100.
However, one of the most important roles played by the user manual is to help in solving problems with Teledyne THCD-100. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Teledyne THCD-100 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center