page 2 Manual Print History Manual Print History Manual Print History Manual Print History Manual Print History The pr int histor y shown below lists the printing dates of all revisions and addenda created for this manual. The revision level letter increases alphabetically as the manual undergoes subsequent updates .
page 3 T T T T T able of Contents able of Contents able of Contents able of Contents able of Contents 1.0 1.0 1.0 1.0 1.0 GENERAL INFORMA GENERAL INFORMA GENERAL INFORMA GENERAL INFORMA GENERAL INFORMA TION TION TION TION TION ........................
page 4.
page 5 General Information SECTION 1 The Hastings Mass Flowmeter, HFM-60 is designed to accurately measure mass flow up to 70 slpm with an accuracy of better than 3%. Hastings mass flow instruments do not require any periodic maintenance under normal operating conditions with clean gases.
page 6 1.2 Specifications Accuracy ......................................................................................................................... <±3% F .S. Repeatability .................................................................
page 7 1.3 Optional 4-20 mA Current Output An option to the standard 0-5 VDC output is the 4-20 mA current output that is propor tional to flow . The 4 - 20 mA signal is produced from the 0 - 5 VDC output of the flowmeter.
page 8.
page 9 Installation SECTION 2 This section contains the necessar y steps to assist in getting a new flowmeter in operation as quickly and easily as possible.
page 10 There is a threaded hole in the bottom of the transducer that can be used to secure it to a mount- ing bracket, if desired. Other holes for special mounting can be added to the end cap as desired.
page 11 2.6.4 Range Changes: The range of the flowmeter can be changed in the field if recalibra tion facilities are a vailable. The flowmeter ma y require a different shunt which can be purchased from the factor y .
page 12.
page 13 Theory of Operation SECTION 3 This section contains an overall functional description of the HFM-60. Detailed schematics and parts lists can be found at the end of the manual in Section 5.
page 14 3.3 Base The 6061 aluminum alloy base has a 1" square cross-section and is 3.375" long (with end cap). The internal flow channel is 0.812" diameter with in and out 304L stainless steel fittings 1/4" NPT to 1/4" Swagelok fittings.
page 15 Aluminum epoxy figure 3.3 (a) 10 sccm - 200 sccm (typical configuration) figure 3.3 (b) annular flow passages 500 sccm - 70 slm (typical configuration) flow flow.
page 16 3.5 Electronic Circuitr y The HFM-60 employs a thermal flow sensor (capillar y tube descr ibed in section 3.2) to measure the flow which is proportional to the total flow through the instr ument. The sensor develops a differential voltage output signal proportional to flow which is approximately 1 mV full scale magnitude.
page 17 4.1 Authorized Maintenance With proper care in installation and use, the flowmeter will require little or no maintenance. If maintenance does become necessar y , most of the instr ument can be cleaned or repaired in the field. Some procedures may require recalibration.
page 18 4.3 Adjustments Calibration Procedure (See figure 4.1) 1. Connect pow er to pins 1 and 2 as specified in Section 2.5. Allow the instrument to warm up for 30 minutes with 10% flow .
page 19 4.4 End Cap Removal The end cap on the inlet side must be remov ed to gain access to the filter and shunt assembly . First remove power and shut off the supply of gas to the instrument. Disconnect the Swagelok fittings on the inlet and outlet sides of the transducer and remove it from the system plumbing.
page 20 4.5 Printed Circuit Board Replacement In the event that any of the PC boards fail, they are easily removed from the instrument and replaced with a spare. This ease in disassembly and replacement substantially reduces instrument downtime. 1 . Replacement of the 4-20 mA option PC board: Unplug the power cable from the trans ducer.
page 21 W arranty and Repair SECTION 5 5.1 W ar ranty Repair P olic y Hastings Instruments warrants this product for a period of one year from the date of shipment to be free from defects in material and workmanship . This warranty does not apply to defects or failures resulting from unauthorized modification, misuse or mishandling of the product.
page 22.
page 23 Diagrams and Drawings SECTION 6.
page 24.
page 25.
page 26.
page 27.
page 28.
page 29.
page 30.
An important point after buying a device Teledyne HFM-60 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Teledyne HFM-60 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Teledyne HFM-60 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Teledyne HFM-60 you will learn all the available features of the product, as well as information on its operation. The information that you get Teledyne HFM-60 will certainly help you make a decision on the purchase.
If you already are a holder of Teledyne HFM-60, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Teledyne HFM-60.
However, one of the most important roles played by the user manual is to help in solving problems with Teledyne HFM-60. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Teledyne HFM-60 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center