Instruction/ maintenance manual of the product 401 Teledyne
Go to page of 114
INSTRUCTION MANUAL MODEL 401 O 3 CALIBRATOR © TELEDYNE INSTRUMENTS ADVANCED POLLUTION INSTRUMENTATION DIVISION (T-API) 9480 CARROLL PARK DRIVE SAN DIEGO, CA 92121-5201 TOLL-FREE: 800-324-5190 FAX: 858-657-9816 TEL: 858-657-9800 E-MAIL: api-sales@teledyne.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 SAFETY MESSAGES Your safety and the safety of others is very important. We have provided m any important safety messages in this manual. Please read these m essages carefully.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 TABLE OF CONTENTS SAFETY M ESSAGES .......................................................................................... II TABLE OF CO NTENTS ..................
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 2.1.2 O 3 Analog Output ................................................................................................................. ... 2-1 2.1.3 E 2 ROM Backup Of Software Configuration .
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 5.2.1 D/A Output ............................................................................................................... ............... 5-2 6 HANDLING W ARNINGS .
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 9.6.4 Checking The DC Power Supply Board................................................................................ 9-13 9.6.5 Checking The Pneumatic Sensor Board .
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 LIST OF FIGURES F IGURE 1-1: M ODEL 401 O ZONE C ALIBRATOR B LOCK D IAGRAM .......................................................... 1-11 F IGURE 1-2: M401 R EAR P ANEL E LECTRICAL C ONNECTIONS .
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 LIST OF TABLES T ABLE 1-1: T EST V ALUES ....................................................................................................................... 1-21 T ABLE 2-1: P ASSWORD L EVELS .
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 1 INTRODUCTION 1.1 Preface Teledyne API is pleased that you have purchased the Model 401.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 The warranties set forth in this section shall be of no force or effect with respect to any Product: (i) that has b.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 NOTE Model 401 Photometric Calibrator is designed to be a dedicated calibrator. The M401 Photometer should not be used as a monitor. The Beer-Lambert equation, shown below, calculates the concentration of ozone from the ratio of light intensities.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 The intensity of light is converted into a voltage by the detector/preamp module. The voltage is converted into a number by a voltage-to-frequency (V/F) converter capable of 80,000 count resolution.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 1. 4 Specifications Photometer Ranges User selectable to any full scale range from 100 ppb to 20,000 ppb Zero Noise < 0.3 ppb (rms) Span Noise < 0.5% of reading (rms) Lower Detectable Limit < 0.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 1.4.1 Noise (PPB RMS) The standard deviation of reported concentration calculated over 25 samples times 7 runs minimum , N-1 weighted.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 1.4.10 Flow Rates 800 scc/min ± 10% (standard cubic centimeters/m inute) sample flow rate. 2-5 l/min ± 10% (standard cubic liter/minute) O 3 generator flow rate. 1.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 NOTE See Figure 1-4 for rear panel pneumatic connections. Span gas should only come into contact with PTFE or glass. Leak check all fittings with soap solution. Maximum pressure for leak check is 15 PSIG.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 10. Turn on the M401 by switching the switch on the lower right corner of the front panel (See Figure 2-1). The front panel display should light with a sequence of messages,-API - M401 - software version number, then a normal display as shown in Figure 2-2.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 14. Generate the Zero Air. a. Press ZERO from the main m enu and enter the password (512). b. Press STBY to terminate or press O 3 GN to continue to generate O 3 span gas.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Figure 1-1: Model 401 Ozone Calibrator Block Diagram 1-11.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 8.7 Figure 1-2: M401 Rear Pa nel Electrical Connections 1-12.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Figure 1-3: Rear Panel Electrical a nd Pneumatic Connections (CE Version) 1-13.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 1.6 Electrical And Pneumatic Connections 1.6.1 Electrical Connections Output #1 Ozone concentration - chart recorde.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Figure 1-4: RS-232 Pin Assignments 1-15.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 1.6.3 Pneumatic System The Model 401 Photometric Calibrator is equipped with a compression pum p to deliver up to 5 L/MIN for ozone generator system. It is also equipped with a second pump capable of pulling 800 cc/min across a critical flow orifice.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Figure 1-5: M401 Calibrator Pneumatic Connections 1-17.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Figure 1-6: M401 Rear Panel (External Dry Zero Air) 1-18.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Figure 1-7: M401 Calibrator Chassis Layout 1-19.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 1.7 Operation Verification The Model 401 Calibrator is now ready for operation. 1. Read Section 1.3 and all of Section 4 of the manual to understand the Calibrator operation.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 1.7.1 Final Test And Calibration Values Table 1-1: Test Values Test Parameter Displayed Value Units TIME HH:MM:SS D.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 1.8 Options 1.8.1 Rack Mount With Slides This option, including slides and rack mounting ears, permits the Analyzer to be m ounted in a standard 19" wide x 30" deep RETMA rack.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 2 OPERATION 2.1 Key Features The important features of the Teledyne API Model 401 Photometric Calibrator are listed below.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 2.1.5 Data Acquisition The Teledyne API Photometric Calibrator contains a built-in data acquisition system which keeps track of the average O 3 readings and the last 100 averages.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 2-1: Password Levels Password Level Functions Allowed No password required 0 TEST, MSG, CLR, SETUP-DAS-VIEW O.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 2.2 Front Panel This section describes the operator interface from th e point of view of the front panel. The front panel consists of a 2-line by 40-character alphanumeric display, 8 pushbuttons, and 3 status LEDs.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 2-2: System Modes Mode Meaning STBY xxxx (1) Standby mode (Photometer is operating normally) ZERO CAL x (2) Z.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 2-3: Test Measurements Test Message Meaning TIME=xx:xx:xx Current time-of-day (HH:MM:SS) O 3 SET =xxxx PPB Oz.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 2-4: Warning Messages Warning Message Meaning SYSTEM RESET Issued whenever Analyzer is powered on RAM INITIAL.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Figure 2-2: Illustration Of Normal Display If TEST is pushed, the upper center display cycles through the menu of test parameters, e.g. Sample flow (see Table 2-3). If STBY is pushed, the calibrator is in stand-by mode.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 2.2.3 Status LEDs The three status LEDs to the right of the display indicate the general status of the Model 401 Photometric Calibrator. The green SAMPLE LED indicates the sampling status.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 INTENTIONALLY BLANK 2-10.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 3 CALIBRATION The M401 is designed to be used as a dedicated ozone calibrator with built-in photometer. Zero/Span gas generation and calibration of Model 401 are discussed in Section 3.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 3.3 Automatic Zero/Span (or Multipoint) Check Automatic zero/span gas supply must be enabled in the setup m ode.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 3. 5 Summary Of Front Panel Calibration Controls The calibration controls are summarized below in term s of the button sequences used to access them.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 3. Allow to sample zero gas for approximately 10 m inutes. 4. Follow sequence of SETUP, BCAL, and ZERO.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 NOTE Teledyne API recommends that contact closures remain closed at least 10 minutes to allow for an accurate average zero or span value to be established.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 INTENTIONALLY BLANK 3-6.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 4 SETUP MODE This section describes the setup variables which are used to configure the analyzer.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 4.4 Examining The Ozone Formula Slope And Offset The slope and offset parameters can be examined (or changed) by SETUP-MISC-O3-SLPE or SETUP-MISC-O3-OFFS. The slope and offset parameters are set only during zero and span calibration routines.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 4.5 Setting the O 3 Generation Mode There are three modes of feedback operation that can be selected for the ozone generator. The mode can be changed by pressing SETUP-MISC-O3-GEN-MODE .
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 4.8 Adjusting The Clock Speed In order to compensate for clocks which run a little bit fast or slow, there is a variable called SETUP-MISC-CLK-ADJ . This variable is set to the number of seconds per day by which to speed up or slow down the clock.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 4.12 Setting The Analyzer I.D. Each Analyzer may be programm ed with a unique I.D. number which appears on all RS-232 messages. To set the Analyzer I.D., press SETUP-COMM-ID .
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 4.15 Software Configuration The software configuration can be displa yed by entering the button sequence SETUP-CFG- LIST.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 4-1: Setup Variables Button Sequence Function Default Limits SETUP-EVNT-WAIT Zero cal. duration Span cal. duration 15 min. 15 min. 1-20 min. 1-20 min. SETUP-EVNT-TIME Cal.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 INTENTIONALLY BLANK 4-8.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 5 DIAGNOSTICS The Teledyne API ozone analyzer contains two levels of diagnostics: test measurements which can be viewed at all times (except when in setup) by pressing TEST , and lower level diagnostic operations which can only be performed by pressing SETUP-DIAG .
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 5-1: Test Channel Output Test # Name Purpose 1 NONE No output 2 O3 DETECT Outputs O 3 detector 3 REF DETECT Outputs IZS ref. det. 4 SAMP PRESS Outputs sample press.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 6 HANDLING WARNINGS When a system warning occurs, a warning m essage is displayed and the FAULT LED blinks. A warning indicates that something in the system needs to be checked or adjusted.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 INTENTIONALLY BLANK 6-2.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 7 RS-232 COMMUNICATIONS The Model 401 Photometric Calibrator features a powerful RS-232 interface which is used both for reporting test results and for controlling the Photometer from a host com puter.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 The "X" indicates the message type as shown above in Table 7-1 and "COMMAND" is the command type, each of which is described individually below. The "<CRLF>" is used to terminate the comm and.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 7.2 Warnings Whenever a warning message is displayed on the display, it is also sent to the RS-232 output.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 7-2: Warning Message Clear Commands Command Warning Message Cleared "W WSYSRES<CRLF>" SYSTEM .
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 To do a remote adjustment via the RS-232 interface, the host com puter should issue a message with the following format: " C COMMAND<CRLF> " The commands are sum marized in Table 7-4.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 7.4 Diagnostics The diagnostics mode can be entered from the RS-232 port as well as from the front panel. The diagnostics commands available are listed below. NOTE The diagnostics mode may only be entered via the RS-232 port when the calibrator is in stand by mode.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 7-6: Special Diagnostic Commands Command Function "D SYS-RESET<CRLF>" Reset Calibrator (same .
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 7-8: Test Measurement Request Commands Command Test Measurement "?<CRLF>" RS-232 HELP screen .
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 For example, to see the analyzer lamp tem perature set point, the host computer would issue the command: "V AL.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 7-9: RS-232 Variable Names Var. Name Button Sequence Legal Values AUTO-TIME SETUP-IZSC-TIME 00:00-23:59 AUTO-.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 8 ADJUSTMENTS NOTE The operations outlined in this chapter are to be performed by qualified maintenance personnel only. All adjustments to the Model 401 are easy to make.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Figure 8-1: Electrical Block Diagram 8-2.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 8.2 A/D - D/A Calibration Procedure Due to the stability of modern electronics, this procedure should not have to be performed m ore than once a year or whenever a major sub-assembly is exchanged or whenever analog output voltage range is changed .
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Next, to automatically calibrate all the DACs, press SETUP-MISC-D/A-CAL-DAC.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Figure 8-2: Flow And Pressure Readout Adjustment 8-5.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 8.5 DC Power Supply Overall performance of the DC power supply may be checked by observing the value displayed during test DCPS.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 8.7 Status Lines Table 8-1: The Calibrator Has The Following Status Conditions Output # Pin Pair High/Low Status Co.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 CAUTION UV light present. Do not pull the lamp from the optical bench assembly.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 8.10 Ozone Generator Calibration The ozone generator can be calibrated against the analyzer calibration by using the analytical section of the M401 to determine the ozone generator's output.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 To view the current dark offset, press SETU P-MISC-DARK-VAL. Press EXIT when finished. No password is required to view the dark offset, only to change it. 8.12 Bench Feedback Option The Bench Feedback Option uses the Photometer to control the lamp drive current.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 9 TROUBLESHOOTING NOTE The operations outlined in this chapter are to be performed by qualified maintenance personnel only.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 If these actions all occur, it is probable that the Calibrator's Power Supplies, CPU, and Display are working properly. If any of these actions fail to occur, power and/or CPU operation should be checked as follows.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 9.2.2 Checking The CPU And Display When the analyzer is turned on, the front panel display should energize and the green "Sample" LED should light. If proper DC power is present (see Section 9.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 9-1: Warning Messages Warning Message Meaning Corrective Action ANA LAMP WARNING The O 3 REF value is greater than 5000 mV or less than 2500 mV. Check and adjust Source Lamp and UV detector as described in Section 9.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 9.4 Troubleshooting Using Test Function Values The Model 401 Photometric Calibrator provides the capability to display, on operator demand, the values of Test Functions which allow the observation of key analyzer operating parameters.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 9-2: Test Function Values (Continued) Test Function Meaning Acceptable Values Corrective Action for Unacceptable Values SAMPLE FLOW Sam ple mass flow rate. 700-900 cc.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 9.5.1 Noisy Or Unstable Readings At Zero 1. Check for leaks in the pneumatic system as described in Section 10.3. 2. Confirm that the Zero gas is free of Ozone. 3. Confirm that the Source Lamp is fully inserted and that the lam p hold-down thumb- screw is tight.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 9.5.4 Analog Outputs Do Not Agree With Front Panel Readings 1. Confirm that the DAC offset (SETU P-MISC-D/A-OFFS) is set to zero. 2. Perform a DAC/ADC adjustment and Dark Si gnal adjustment by following the procedure described in Sections 8.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 9.6 Troubleshooting Individual Sub-Assemblies And Components The following sections provide troubleshooting/check-out methods for the specific sub- assemblies and components of the Calibrator.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 9.6.2 Troubleshooting Temperature Problems The Model 401 has been designed to operate at ambient temperatures between 5 o C and 40 o C.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Chassis Temperature Turn the analyzer off and remove the DC Power Supply Board.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 2. Confirm that all jumpers on the V/F board are set properly, as follows: Table 9-3: V/F Board Jumpers - Factory S.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 9.6.4 Checking The DC Power Supply Board A schematic and physical diagram of the DC Power Supply Board are shown on Drawings 00015 and 00016 in Appendix D.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Four Temperature linearization circuits are contained on the DC Power Supply board. The outputs of these circuits can be checked by measuring the voltages at test points on the board as follows: TP1 Sample Temp 30 ° C=2.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 9-5: UV Source Lamp And Detector Diagnostics UV Source Lamp and Detector Diagnostics O 3 REF Value Meaning 4700 mV to 5000 mV The Source Lamp and Detector are operating, but adjustment is required.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 NOTE In cold, ambient conditions, it may require 5 to 15 minutes of warm- up before the source lamp initially fires. Be sure to wait for this period before troubleshooting the lamp/detector.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 9-17 WARNING Hazardous voltage present - use caution. 1. Unplug the cable connector at J1 on the Lamp Power Supply and confirm that +24VDC is present between Pins 1 and 2 on the cable connector.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Warranty/Repair Questionnaire TELEDYN E I NSTRUMENTS Advanced Pol l ut i on I ns t r um ent at i o n Model 401 A Tel edyne Technol ogi es Com pany 9.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 10 ROUTINE MAINTENANCE NOTE The operations outlined in this chapter are to be performed by qualified maintenance personnel only. 10.1 Maintenance Schedule Table 10-1 shows a typical maintenance schedule for the Model 401.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 10.2 Leak Checking There are two methods of leak checking: 1. By vacuum: This is the simplest m ethod but it does not show the location of a leak. 2. By pressure: By using bubble solution, this method shows a leak location.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 10.4 Changing The Prom 1. Locate the CPU card by referring to Figure 1-7. 2. Remove the screws that hold the CPU card (SBC40 printed on the lever) top corner to the V/F card, then remove the card from the STD-BUS backplane.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 INTENTIONALLY BLANK 10-4.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 11 MODEL 401 SPARE PARTS AND EXPENDABLES KITS 11.1 Spare Parts For CE Mark Units NOTE Use of replacement parts other than those supplied by Teledyne API may result in non-compliance with European Standard EN 61010-1.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 11-1: Spare Parts For CE Mark Units (Continued) Part No. Description 01112 M 401 Expendables Kit 01509 UV Lam.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 11.2 Spare Parts For Non-CE Mark Units Table 11-2: Spare Parts For Non-CE Mark Units Part No.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 11-2: Spare Parts For N on-CE Mark Units (Continued) Part No.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Appendix A LIST OF AVAILABLE MODEL 401 OPTIONS Table A-1: Available Model 401 Options Option 1.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 INTENTIONALLY BLANK A-2.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Appendix B RECOMMENDED ZERO AIR SYSTEM The following is from the TECHNICAL ASSISTANCE DOCUMENT FOR THE CALIBRATION OF AMBIENT OZONE MONITORS - Richard J. Paur and Frank F.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 INTENTIONALLY BLANK B-2.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Appendix C REFERENCES 1. Calibration of Ozone Reference Methods, Code of Federal Regulations , Title 40, Part 50, Appendix D.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 INTENTIONALLY BLANK C-2.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 D-1 Appendix D SOFTWARE MENU TREES Figure A-1: Model 401 Sample Software Menu Tree.
ne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 D-2 Figure A-2: Model 401 Setup Software Menu Tree Teledy.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Appendix E ELECTRICAL SCHEMATIC INDEX Table E-1: Electrical Schematic Index Part No.
Teledyne API Model 401 O 3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 INTENTIONALLY BLANK E-2.
An important point after buying a device Teledyne 401 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Teledyne 401 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Teledyne 401 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Teledyne 401 you will learn all the available features of the product, as well as information on its operation. The information that you get Teledyne 401 will certainly help you make a decision on the purchase.
If you already are a holder of Teledyne 401, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Teledyne 401.
However, one of the most important roles played by the user manual is to help in solving problems with Teledyne 401. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Teledyne 401 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center