Instruction/ maintenance manual of the product LC 500 Siemens
Go to page of 122
Instruction Manual January 20 03 LC 50 0 sitr ans.
© Siemens M i l l troni c s Pro c e s s In s tr u ments In c . 2 00 3 Safety Guid elines Warning not ices must be observ ed to ensure per sonal safety as well as t hat of other s, and to protect t he product and the connect ed equipment. These war ning notices ar e accompanied by a clarificat ion of t he level of caution to be observed.
Ta b l e o f C o n t e n t s i mmmmm T able of Contents SITRANS LC 50 0 .............. .. ........... .......... .......... .. ........... .......... .......... .. ........... .......... ......... .. 1 Applications ....................................
ii mmmmm Ta b l e o f C o n t e n t s Cathodically Prot ected Metal Tanks ........... .................... ..................... .................. ..............29 Non-Conductive Tan ks .......... .................... .................. .............
iii mmmmm T able of Contents Appendix C: HAR T Documenta tion .......... .......... ........... .......... ........... .......... .......... .. ..... 75 HART Communic ations for the SITRANS LC 50 0 .............. ......................................
iv mmmmm Ta b l e o f C o n t e n t s Appendix F : Approvals . .......... ........... .......... ............. .......... .......... ........... .......... .......... ..... 96 CE Cert ificate ............. ..................................... .......
7ML19985GE0 1 SITRANS LC 50 0 – INSTRUCTION MANU AL P age 1 mmmmm Introduction SITRANS LC 50 0 SITRANS LC 50 0 is a high performance 2-wir e capacitance instrument for continuous level and inter face measur ement in extreme or critical conditions.
P age 2 SITRANS LC 50 0 – INSTRUCTION MANU AL 7ML19985GE0 1 mmmmm Introduction Safe t y N o t es Special attent ion must be paid to warnings and not es highlighted from the rest of the text by gr ey boxes. Saf ety marking symbols The M a n u a l This manual will help you set up y our SI TR ANS LC 50 0 for opt imum per formance.
7ML19985GE0 1 SITRANS LC 50 0 – INSTRUCTION MANU AL P age 3 mmmmm Introduction Abbreviations and Identifications Short form Long Form Des cr ip t io n Units A/D Analog to Digital CE / FM / CSA Confo.
P age 4 SITRANS LC 50 0 – INS TRUCTION MANU AL 7ML19985GE01 mmmmm Specifications T ec hnical Specifications: SITRANS LC 50 0 Po w e r Supply v olt age • m aximum: 33 Vdc, (30 V d c for I S ) • minimum 12 Vdc at 3.6 mA (9.5 Vdc at 22 mA) Loop current 3.
7ML19985GE0 1 SITR ANS LC 50 0 – INS TRU CTION MANU AL P age 5 mmmmm Specifications Safety • current signalling ac cor ding to NAMUR NE 43; 3.6 to 22 mA / 22 to 3.
P age 6 SITRANS LC 50 0 – INS TRUCTION MANU AL 7ML19985GE01 mmmmm Specifications Electrodes Pr ocess connections • thr e aded connection AISI 31 6 L st ainless steel, 3/4” , 1” , 1-1/4” , 1 -1/2” , 2” , NPT , BSPT , JIS • flat-faced flanges AISI 316 L stainl ess steel (optiona l C 22.
7ML19985GE0 1 SITR ANS LC 50 0 – INS TRU CTION MANU AL P age 7 mmmmm Specifications Pr ocess Condit ions Pr essure rating 1 • standard FV (full v acuum) to 20 0 bar (2920 psi) • option up to 525.
P age 8 S ITRANS LC 50 0 – INS T RUCTION MANUAL 7ML19985GE0 1 mmmmm Operation & Application SITRANS LC 50 0: T ransmitter Operating P rinciples Capacit ance 1 measurement oper at es by forming a variable c a pacitor resulting from the installat ion of a vert ical measur ement electr ode in a vessel or silo.
7ML19985GE0 1 SITRANS LC 50 0 – INSTRUCTION MANU AL P age 9 mmmmm Operation & Application The formula 1 is: where C = capacitance K = constant ε = dielectric constant L = active measurement length D = diameter of t ank d = probe diameter . (For detailed applic atio n examp les, see page 91.
P age 10 SITRANS LC 50 0 – INS TRUCTION MANUAL 7ML19985GE01 mmmmm Operation & Application In a non-conductive or irr egular t ank Where the v essel contents are non-conductive: • a referenc e .
7ML19985GE0 1 SITRANS LC 50 0 – INSTRUCTION MANU AL P age 1 1 mmmmm Operation & Application The meas urement is further protected from interferenc e by a buffer , whic h applies the frequenc y signal from the measur ement sect ion to the active shie ld sect ion.
P age 12 SITRANS LC 50 0 – INS TRUCTION MANUAL 7ML19985GE0 1 mmmmm Operation & Application The entire SITRANS LC 50 0 tr ansmitt er is pott ed in epoxy resin as part of the intrinsic safet y protect ion. The potting also protects the electronics against mech anical vibration and moisture influences.
7ML19985GE0 1 SITRANS LC 50 0 – INSTRUCTION MANU AL P age 13 mmmmm Operation & Application 3. Upper T hreshold Setting and Lower T hreshold S etting act ivate and deactivate the 2-state output, .
P age 14 SITRANS LC 50 0 – INS TRUCTION MANUAL 7ML19985GE01 mmmmm Operation & Application Inter face Measurement The c apacit ance of the electrode s yste m is dependent on the dielectric constant of the product surrounding the pr obe.
7ML19985GE0 1 SITRANS LC 50 0 – INSTRUCTION MANU AL P age 15 mmmmm Operation & Application Fault Signalling The SITRANS LC 50 0 has three fault signallin g options: • via the loop-current • via HAR T • via the solid-stat e output or solid-st at e relay .
P age 16 SITRANS LC 50 0 – INSTRUCTION MANU AL 7ML19985GE01 mmmmm Pr obe Configuration SITRANS LC 50 0: P robe Configur at ion The pr obe (electrode) comprises a measuremen t section and an active shiel d sect ion. This electr ode connects to the c apacit ance detector por tion of the t wo-wire loop powered electronic tr ansmitter .
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MANUA L P age 17 mmmmm Pr obe Configuration Electrode Assembly SITRANS LC 50 0 electrodes c ome in a variet y of formats to provide the necessary char acterist ics for corr ect mount ing, chemi c al compatibility , t emperatur e and pressure requir ements, and dielectric constant of the medium.
P age 18 SITRANS LC 50 0 – INSTRUCTION MANU AL 7ML19985GE01 mmmmm Pr obe Configuration P ressure and T emperature Considerations The maximum temperature and pr essure of oper at ion for the standard SITRANS LC 50 0 lev el probe is 20 0 ° C (392 ° F) and 20 0 bar (290 0 psi).
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MANUAL P age 19 mmmmm Pr obe Configuration Non-standard applications Applications outside the standard c apabilities of the S-Series r equire a different design configuration.
P age 20 SITRANS LC 50 0 – INS TRUCTION MANUAL 7ML19985GE0 1 mmmmm Installation SITRANS LC 50 0: Installat ion Handling Electrodes Notes: • Installation shall only be per formed b y qu alif ied per sonnel and in accor dance with local gov erning regulations.
7ML19985GE0 1 SITRANS LC 50 0 – INS TR UCTION MANU AL P age 21 mmmmm Installation Mounting Instructions The SITRANS LC 50 0 is easily inst alled: s imply mount the instru ment on the proc e ss conn ect ion of the vessel .
P age 22 SITRANS LC 50 0 – INS TRUCTION MANUAL 7ML19985GE0 1 mmmmm Installation SITRANS LC 50 0: St andard Lev el V ersion Av ailable with the following features: • Thr eaded f langes, welded fl a.
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MANUAL P age 23 mmmmm Interconnection Interconnection: SITRANS LC 50 0 Wiring Supply The SITRANS LC 50 0 uses a switched powe r supply circuit, which makes the most efficient use of the available power present on t he t erminals.
P age 24 SITRANS LC 50 0 – INS T RUCTION MANU AL 7ML1 9985GE01 mmmmm Interconnection Cable Selecting the corr ect instrument ation cable • you need to know the cable length, the barrier t ype (if applicable), and the measurement resistanc e • select a cable that will giv e you a capacita nc e t ime c onst ant of less than 65 µS ec 1 .
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MANUAL P age 25 mmmmm Interconnection 2. Calculat e the length of cable allowed, b y subtr acting the capacitance valu e that the devic e presents on the loop from the total capacitance. SITRANS LC 50 0 has no measurable c apacitance value, but assume 5 nF .
P age 26 SITRANS LC 50 0 – INS T RUCTION MANU AL 7ML1 9985GE01 mmmmm Interconnection Te r m i n a l s The SITRANS LC 50 0 is equipped with t wo te rminal blocks, both inse nsitive to polarit y . One terminal block is int ended for conn ecting the instrument cable (loop power).
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MANUAL P age 27 mmmmm Interconnection P rotect ion for solid-sta te switc h • For dc cir cuits: connect protection diodes in the the correct polarity across the relay coil.
P age 28 SITRANS LC 50 0 – INS T RUCTION MANU AL 7ML1 9985GE01 mmmmm Interconnection Grounding Ex amples: SITRANS LC 50 0 Groundin g is impor tant for t wo reasons: 1 . T o pre vent int erference to the signal: s yst em grounding 2. For safet y purposes: safet y grounding Se veral c ommon applications are illustr ated.
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MANUAL P age 29 mmmmm Interconnection Note: Grounding the SITRANS LC 50 0 as illu strated abov e provides only sy st em grou nding for referencin g purposes: it does not provide safety grounding.
P age 30 SITRANS LC 50 0 – INS T RUCTION MANU AL 7ML1 9985GE01 mmmmm Interconnection Saf ety Grounding The safety grounding r equirements are deter mined b y the application and the connected instruments.
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MANUAL P age 31 mmmmm Interconnection Example 3 If you do not want to ground the negative connection directly , or in the case of Intrinsically S afe app.
P age 32 SITRANS LC 50 0 – INS T RUCTION MANU AL 7ML1 9985GE01 mmmmm Interconnection Communications The SITRANS LC 50 0 is equipped with HART c ommunication 1 protocol 2 so that settings and values can be obtained and altered locally 3 or remotely .
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MANUAL P age 33 mmmmm Interconnection Whenever the loc al sit uation allows, the zero adjustment and the full scale c an be set using the push-button feat ure and the appropriate menu select ion.
P age 34 SITRANS LC 50 0 – INS T RUCTION MANU AL 7ML1 9985GE01 mmmmm Interconnection Upper and Lower T hreshold, (menus 1 3 and 14) and the a ccompanied delay s (menus 15 and 16). Initially the solid-st ate output is disabled for both signal ou tput and fault/failure output (menu 17).
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MANUAL P age 35 mmmmm Interconnection • The Upper S ensor Limit (USL) and Lower Sensor Limit (L SL) are set to 330 0 and 1 .
P age 36 SITRANS LC 50 0 – INSTRUCTION MANU AL 7ML19985GE0 1 mmmmm User Interface T he SITRANS LC 50 0 User Int er face The SITRANS LC 50 0 user inter face c onsists of the display (LCD), the rotary switch, and two push-buttons.
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MANU AL P a ge 37 mmmmm User Inter face How to access the dat a: Ac ces s data in th e tr ansm itter fro m 28 menu items divided bet ween two menu levels: 00 to 0F and 10 to 1F . Use the rotary s witch and push-buttons in c ombinat ion to select an item and adjust the value.
P age 38 SITRANS LC 50 0 – INSTRUCTION MANU AL 7ML19985GE0 1 mmmmm User Interface Th e ro ta ry sw i t ch The r ot ary switch giv es y o u access first to the menu level and then to the menu item. The r ot ary switch has a small slot where the current position can be read.
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MANU AL P a ge 39 mmmmm User Inter face T ransmitter V ariables • T ransmitter V ariable 0 (TV0) is the c apacit ance as measur ed by the device. • T ra nsmitt er V ariable 1 (TV1) is a c omp ut ed variable: t he dynamic value is a computed derivative from the r ange settings for TV0.
P age 40 SITRANS LC 50 0 – INS TRUCTION MANUAL 7ML19985 GE01 mmmmm Start Up Start-up: SITRANS LC 50 0 Capacitance measurement syst ems requir e t he instrument to be calibr at ed for a part icular application. T wo metho ds of calibr ation are available: • push-button (for instructions, see page 43).
Rotary Switc h Po sit ions – Quick Ref erence 01 2 3 4 5 6 7 8 9 A B C D E F Menu LEVEL 0 (0 0 to OF) Units PV Units Num. Selection PV Units Se con ds Se con ds % % mA mA Numeri- cal Numeri- cal PV .
P age 42 SITRANS LC 50 0 – INS TRUCTION MANUAL 7ML19985 GE01 mmmmm Start Up Menu levels 0 and 1 Menu level 0 0 to 0F is the default st art-up s ettin g after power is applied or after a reset. Menu Leve l 10 through 1F is fl agged in the LC D b y an left-arrow indicator in the upper left corner of the LCD .
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MANUAL P age 43 mmmmm Start Up Calibration using push-button adjustment Reset keyloc k level if necessary t o enable sett ings to be changed: menu 1F (no change is nec essary if the factory se tting has not been changed) 1 .
P age 44 SITRANS LC 50 0 – INS TRUCTION MANUAL 7ML19985 GE01 mmmmm Start Up Method 2. Set v alue for 10 0% (URV): v alues must be displayed as per cent (at menu 01 , Pv = P) a. Calculate the percentage v alue of the current level. b. Set r ot ary switch to 1 , and pr e ss both buttons to adjust the value to P .
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MANUAL P age 45 mmmmm Start Up Se lect capacitance or per cent mode for dynamic prim ary variable (PV): menu 01 a. T o see the value displa yed as per cen t, select menu 01 and press both buttons to set Pv = P .
P age 46 SITRANS LC 50 0 – INS TRUCTION MANUAL 7ML19985 GE01 mmmmm Start Up Calibration using HART The SITRANS LC 50 0 transmitter can be calibr ated using HART , with a HART com m un ica t or 1 ; a laptop running Simatic PDM, or wi th the Host sy st em (DCS).
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MANUAL P age 47 mmmmm Start Up Example 2 For situations where the capacitance values are known in adv a nce. 1 . Switch on the 275 and establish co nnection with the SITR ANS LC 50 0. a. Select: Online b.
P age 48 SITRANS LC 50 0 – INS TRUCTION MANUAL 7ML19985 GE01 mmmmm Start Up The c apacitance value for a 17% c hange in level is 17 * 2.08 = 35.37 pF . The c apacitance value for 0% is 52 – 35.37 = 16.62 pF (initial capacitance v alue). The c apacit ance v alue for 10 0% is (10 0 * 2.
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MANUAL P age 49 mmmmm Start Up (130) Read Failsafe Mode selection (131) Ret ur n dev ice con f ig ura ti on i nfo (132) S et V ariable Upper Limit (133) .
P age 50 SITRANS LC 50 0 – INS TRUCTION MANUAL 7ML19985 GE01 mmmmm Maintenance Maintenance T est function Auto Sel f-testi ng SITRANS LC 50 0 c o ntinuously per forms a variety of tests to verify that the device is functioning correctly . T hese include a test whe re a known c apaci tor is applied to the input of the devic e.
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MANUAL P age 51 mmmmm Maintenance f . T he coaxial connector is free of dirt or deposits. g. No cables or wir es are jammed under the co ver . 2. Function al Checks a. Check for r equired minimum terminal vo ltage (see page 23 for supply voltage requir ements).
P age 52 SITR ANS LC 50 0 – INS TRUCTION MANUAL 7ML19985GE01 mmmmm T roubl eshooting T roubleshooting: SITRANS LC 50 0 If you ar e unable to change settings: • chec k that keylock le vel (menu 1F).
7ML19985GE0 1 SITRANS LC 50 0 – INSTRUCTION MANUAL P age 53 mmmmm T roubl eshooting Error Messag es and Error Codes Error Messag es (push-butt on operation) Error Codes (HART) Error Message Des cr ip ti on Cause Flt a a. Alternat es with the primary variable (PV).
P age 54 SITRANS LC 50 0 – INS TRUCTION MANUAL 7ML19985GE01 mmmmm Appendix A Appendix A: Menu Groups The data in the transmitter is acces sible as 28 1 menu items divided bet ween t wo menu lev els: 00 to 0F and 10 to 1F . Y ou can switc h bet ween the two levels at position 00 and 10 .
7ML19985GE0 1 SITRANS LC 50 0 – INSTRUCTION MANU AL P age 55 mmmmm Appendix A Menu Items T ransmitter: V ariable Settings: menu level 0 Stepsize Update Value This menu selection contr ols the increment/decrement step-size for the menus 0B, 0C, 0D, 0E, 0 F , and 03.
P age 56 SITRANS LC 50 0 – INS TRUCTION MANUAL 7ML19985GE01 mmmmm Appendix A Dam pi ng Damping slows the m e asurement r esponse to a ch ange in level, and is used to stabilize the reading 1 . The Damping V alue is not in seconds but is a factor that controls the r at e of change for the dynamic value of the TV curr ently selected.
7ML19985GE0 1 SITRANS LC 50 0 – INSTRUCTION MANU AL P age 57 mmmmm Appendix A Upper S ensor Limit The Upper Sensor Limit (USL) is the upper of t wo limit settings. Whenever the PV value (Menu Leve l 0) rises above the upper limit settin g, the measurement is consider ed at fault and the LCD displays ooL , alternating with PV .
P age 58 SITRANS LC 50 0 – INS TRUCTION MANUAL 7ML19985GE01 mmmmm Appendix A The tr ansmitt er variable on which this selection is based is chosen in Menu 01 . The increment/decrement step size is set at Menu 09. Lower Range V alue Lower Range V alue (LRV) is the sett ing for 0% of the operating range, in most cases an empty vessel/t ank.
7ML19985GE0 1 SITRANS LC 50 0 – INSTRUCTION MANU AL P age 59 mmmmm Appendix A Upper Range V alue Upper Range V alue (URV) is the sett ing for 10 0% of the operating ra nge, in most cases a full vessel/tank. The tr ans mitter variable on which this menu selection operates is chosen in Menu 01 .
P age 60 SITRANS LC 50 0 – INS TRUCTION MANUAL 7ML19985GE01 mmmmm Appendix A (SIM) function has been sele cted via HA RT , the LCD alternately displays the text SIM or the applied simulat ion value for t he duration of the simulation. T o change from menu 10 to menu 0 0: 1 .
7ML19985GE0 1 SITRANS LC 50 0 – INSTRUCTION MANU AL P age 61 mmmmm Appendix A 1 . Set the rotary switch to 1 . The LCD displays Pv = 0 , 1 , or P . 2. Press the RED (+) or BLUE (–) button to select a higher or lower v alue. 3. Press both buttons to select Pv = P .
P age 62 SITRANS LC 50 0 – INS TRUCTION MANUAL 7ML19985GE01 mmmmm Appendix A Analog Output Signalling (pro port ional or 2-st ate): menu level 0 Analog mode (the loop-curr ent) can pr ovide either: .
7ML19985GE0 1 SITRANS LC 50 0 – INSTRUCTION MANU AL P age 63 mmmmm Appendix A 1 . Set the loop-current contr ol (Menu 07) to 2-stat e mode. 2. Set the rotary switch to 4 . 3. Press the RED (+) or BLUE(–) buttons to increase or decrease the v a lue.
P age 64 SITRANS LC 50 0 – INS TRUCTION MANUAL 7ML19985GE01 mmmmm Appendix A Lower T hreshold Setting (2-state mode) The Lower Thr eshold Setting is the % of range below which the probe is consider ed unco vered. In order to switc h the output si gnal, the corresponding delay time has to be met (Menu 04).
7ML19985GE0 1 SITRANS LC 50 0 – INSTRUCTION MANU AL P age 65 mmmmm Appendix A Menu selections 03, 04, 05, and 06 s et the criteria for delay and t hreshold that have to be met for a change in output signal.
P age 66 SITRANS LC 50 0 – INS TRUCTION MANUAL 7ML19985GE01 mmmmm Appendix A Analog Fault Signalling (2-state) When 2-stat e fault signalling is enabled, in the case of a fault the mA output is 3.6 mA or 22 mA 1 , depending on the sett ing. The mA output is viewed at menu 07 .
7ML19985GE0 1 SITRANS LC 50 0 – INSTRUCTION MANU AL P age 67 mmmmm Appendix A Digital Output Signalling (solid -state output): menu level 1 T o set values for Upper an d Lower Thre shold Delay , an d Upper and Lower Thr eshold Setting, the solid-stat e switch output must be enab led (menu 17 ).
P age 68 SITRANS LC 50 0 – INS TRUCTION MANUAL 7ML19985GE01 mmmmm Appendix A Lower T hreshold Delay (solid-state output) The Lower Thr eshold Dela y contr ols the Deacti vation delay: the amount of t ime that has to pass uninterrupted with the probe co vere d to a level below the Lower T hreshold Setting before the timer expires.
7ML19985GE0 1 SITRANS LC 50 0 – INSTRUCTION MANU AL P age 69 mmmmm Appendix A First select the solid-stat e switch output at menu 17 ( contact open or co nt act closed ). 1 . Set the rotary switch to 5 . 2. Press the RED (+) or BLUE (–) button to increase or decrease the value.
P age 70 SITRANS LC 50 0 – INS TRUCTION MANUAL 7ML19985GE01 mmmmm Appendix A Digital Signalling Mode (solid-state output) Controls the solid-state switch output and allows you to set the switc h to contact open or contact closed . The settings are relative to a co vered pr obe, and t he criteria are set in menu 15 and 16.
7ML19985GE0 1 SITRANS LC 50 0 – INSTRUCTION MANU AL P age 71 mmmmm Appendix A Digital Fault Signalling Controls the solid-stat e switch response to a fault/failur e and allows you to select either contact open or contact closed . With contact closed the solid-stat e switch will be on: with co nt a c t open the solid-stat e switch will be off .
P age 72 SITRANS LC 50 0 – INS TRUCTION MANUAL 7ML19985GE01 mmmmm Appendix A Miscellaneous Output Signal Pr ocessing T est Displays the Fault/Failure inform ation. If oper at ion is normal, two te st displays alternate, which light up all the LCD segments in a c ycle.
7ML19985GE0 1 SITRANS LC 50 0 – INSTRUCTION MANU AL P age 73 mmmmm Appendix A Range Inversion Displays whether the devic e is ope r at ing with a normal or inverted ra n g e s et t in g . A normal range setting is where LRV (Menu 0E) is lower in va lue than URV (Menu 0F): t he LCD display s nor .
P age 7 4 SITRANS LC 50 0 – INSTRUCTION MANU AL 7ML19985GE01 mmmmm Appendix B Appendix B: LCD display ex amples LCD: alphanumeric display ex amples Menu Item Indicator: Menu Level Indicator: Interna.
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MANUAL P age 75 mmmmm Appendix C Appendix C: HART D ocument at ion HART 1 Communications for the SITRANS LC 50 0 Highway A ddressable Remote T r ansducer (HAR T) is an industrial pro tocol that rides on top of a 4-20 mA signal.
P age 76 SITRANS LC 50 0 – INSTRUCTION MANU AL 7ML19985GE01 mmmmm Appendix C SITRANS LC 50 0 DD Menu/V ariable Organization Device set up menu PV digital va lue PV upper range value PV lowe r range .
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MANUAL P age 77 mmmmm Appendix C HART Response Code Information Additional r esponse code information, S econd Byte. Bit #7: Field Device Malfunction When the tr ansmitt er det ects a malfunction, th e Analog Output will be set in a fault state.
P age 78 SITRANS LC 50 0 – INSTRUCTION MANU AL 7ML19985GE01 mmmmm Appendix C HART Conformance and Command Class SITRANS LC 50 0 transmitter Confor mance and Co mmand Class summary .
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MANUAL P age 79 mmmmm Appendix C General T ransmitter Information Damping information The SITRANS LC 50 0 transmitter implement s damping on most of the transmitter v a r ia b l e s .
P age 80 SITRANS LC 50 0 – INSTRUCTION MANU AL 7ML19985GE01 mmmmm Appendix C MultiDro p operation The SITRANS LC 50 0 transmitter suppor ts MultiDrop Operation. Burst mode The SITRANS LC 50 0 transmitter does not suppor t Burst Mode. Units conversions The T r ansmitter V ariable #0 Units ar e in pF a nd cannot be changed.
Appendix D: Bloc k Diagram, and Corr elat ion table, mA t o % sensor circuit TV 0 ran g e bloc k damping block TV s el e ct (menu 01) max/min reco rde d val ue status 2-state TV1 ran g e block damping.
P age 82 S ITR ANS LC 50 0 – INSTRUCTION MANU AL 7ML19985GE01 mmmmm Appendix D Correlation T able: 0% - 10 0% to 4-20 mA or 20-4 mA Range 0 - 1 0 0 % Current in mA Range 1 0 0 - 0 % 04 . 0 1 0 0 5 4.8 95 10 5.6 90 15 6.4 85 20 7 .2 80 25 8.0 75 30 8.
7ML19985GE0 1 SITRANS LC 50 0 – INS TR UCTION MANUAL P age 83 mmmmm Appendix E Appendix E: SITRANS LC 50 0, alt ernate versions and application details Standard V ersion Standard V ersion S-S eries, Threaded Note: All measurements are giv en in millimet ers/inc hes.
P age 84 SITRANS LC 50 0 – INSTRUCTION MANUAL 7ML1 9985GE01 mmmmm Appendix E Standard V ersion S-S eries, Threaded Features: Sta ndard V ersion S-Series, Thr eaded • single proc e ss seal • suit.
7ML19985GE0 1 SITRANS LC 50 0 – INS TR UCTION MANUAL P age 85 mmmmm Appendix E Standard V ersion S-S eries, Welded and Mac hined Flanged V ersions Features: Standard V ersion S-Series, Fla nged • .
P age 86 SITRANS LC 50 0 – INSTRUCTION MANUAL 7ML1 9985GE01 mmmmm Appendix E Standard V ersion D-S eries, Machined Flange Note: All measurements are giv en in millimet ers/inc hes.
7ML19985GE0 1 SITRANS LC 50 0 – INS TR UCTION MANUAL P age 87 mmmmm Appendix E Note: All measurements are giv en in millimet ers/inc hes. SD-Series Pr obe/Thermal Isolator Features: Standard V ersion SD-Series • double process seal • redundant safety (e.
P age 88 SITRANS LC 50 0 – INSTRUCTION MANUAL 7ML1 9985GE01 mmmmm Appendix E Inter face V ersion Designed specifically for interf ace level where a long dist ance active shield port ion of the electrode is requir ed (up to 35 met ers) befo re the measurement portion of the electrode begins.
7ML19985GE0 1 SITRANS LC 50 0 – INS TR UCTION MANUAL P age 89 mmmmm Appendix E Sanitary V ersion The h ygienic design includes thre aded and tri-clamp versions for use in the food and pharmac eutical industry . Sanitary V ersions, threaded and tri-clamp Note: All measurements are giv en in millimet ers/inc hes.
P age 90 SITRANS LC 50 0 – INSTRUCTION MANUAL 7ML1 9985GE01 mmmmm Appendix E Flanges Flange Standards Notes: • All measurements are giv en in mm • One (1) inc h = 25.4 mm • For det ails, see drawings, technical data, and measuring probe det ails on pages 83 to 89.
7ML19985GE0 1 SITRANS LC 50 0 – INS TR UCTION MANUAL P age 91 mmmmm Appendix E Applications Examples Generic Application Calculations The c apacit anc e expect ed in a cylindric a l tank with a prob.
P age 92 SITRANS LC 50 0 – INSTRUCTION MANUAL 7ML1 9985GE01 mmmmm Appendix E After calibration: 12. 7 p F ≅ 0% ≅ 4 mA or 20 mA 25.4 pF ≅ 10 0% ≅ 20 mA or 4 mA Larger tank, dimensions in fe et: For this slightly larger tank, the capacitance r anges f rom 16.
7ML19985GE0 1 SITRANS LC 50 0 – INS TR UCTION MANUAL P age 93 mmmmm Appendix E Application : level indicator and solid-stat e switc h output The loop current pr ovides: • a reading proport ional t.
P age 94 SITRANS LC 50 0 – INSTRUCTION MANUAL 7ML1 9985GE01 mmmmm Appendix E Application : Analog fault signalling (2-state output) In 2-stat e mode, the loop curr ent signals whether the probe is c overed or unc over ed, and the continuous lev el meas urement is unavailable.
7ML19985GE0 1 SITRANS LC 50 0 – INS TR UCTION MANUAL P age 95 mmmmm Appendix E USL is set to 83.50 pF Menu 0C reads 83.50 LS L is set to 7 .3 pF Menu 0B r eads 7. 3 0 When the proc ess level reac hes 90%, the prob e is consider ed cover ed and the output will be 20 mA.
P age 96 SITRANS LC 50 0 – INSTRUCTION MANUAL 7ML19985GE0 1 mmmmm Appendix F Appendix F: Appro vals CE Certificate WRITTEN DECLAR A TION OF CONFORMIT Y We , Siemens Milltr onics Pr oc ess Instruments B.
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MA NU AL P age 97 mmmmm Appendix F CE Besc heinigung SCHRIFTL ICHE K ONFORMITÄTSERKLÄR UNG Erklärung der Siemens Milltronics Proc e ss I nstrumen ts B.
P age 98 SITRANS LC 50 0 – INSTRUCTION MANUAL 7ML19985GE0 1 mmmmm Appendix F Instrument label: SITRANS LC 50 0.
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MANUAL P age 99 mmmmm Appendix F KEMA certificat e and sc hedules.
P age 10 0 SITRANS LC 50 0 – INSTRUCTION MANUAL 7ML19985GE0 1 mmmmm Appendix F.
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MANUAL P age 10 1 mmmmm Appendix F.
P age 102 SITRANS LC 50 0 – INSTRUCTION MANU AL 7ML19985GE0 1 mmmmm Appendix F.
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MANUAL P age 103 mmmmm Appendix F.
P age 104 SITRANS LC 50 0 – INSTRUCTION MANU AL 7ML19985GE0 1 mmmmm Appendix F.
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MANUAL P age 105 mmmmm Appendix F.
P age 106 SITRANS LC 50 0 – INSTRUCTION MANU AL 7ML19985GE0 1 mmmmm Appendix F.
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MANUAL P age 107 mmmmm Appendix F Certificates and Approv als The Intrinsic S a fety Specif ications of the SITRANS LC 50 0 have been defined and approv ed as follows: The SITRANS LC 50 0 can be directly c onnec t ed to an intrinsic a lly safe supply for intrinsically safe applications .
P age 108 SITRANS LC 50 0 – INSTRUCTION MANU AL 7ML19985GE0 1 mmmmm Appendix F Control Dr awing FM/CSA Approv al SITRANS LC 50 0 MERCAP/FTS MERCAP/N MERCAP/FTF MERCAP/F MERCAP 500 / POINTEK CLS500 UNIT Class I, Div . 1, Groups A, B, C, and D: Vmax = 30 V , Imax = 110 mA, Ci = 0, Li = 10µH Class I, Div .
7ML19985GE0 1 SITRANS LC 50 0 – INSTRUCTION MANU AL P age 109 mmmmm Glossary Glossary cap a ci t a nce : the propert y of a system of conductor s and dielectrics that permits th e stor age of electricity whe n potent ial differences exist betwee n the conductors.
P age 1 10 SITRANS LC 50 0 – INSTRUCTION MANU AL 7ML19985GE0 1 mmmmm Glossary Notes.
7ML19985GE0 1 SITRANS LC 50 0 – INS TRUCTION MANUAL P age 1 1 1 mmmmm Index Index Numerics 2-state output analog fault signalling example 94 A Abbreviations and Identifications list 3 analog fault s.
P age 1 1 2 SITRANS LC 50 0 – INS TRUCTION MANUAL 7ML19985 GE01 mmmmm Index non-conductive tanks system grounding example 29 non-volatile memo ry data storage 79 O one point calibration 33 operating.
7ML19985GE0 1 SITRANS LC 50 0 – INSTRUCTION MANUAL P age 1 13 mmmmm Quick Ref erence Quic k Refer ence: SITRANS LC 50 0 Quic k Start W e strongly rec ommend you re ad the full ma nual to use your device to its fullest potential.
Rotary Switc h Po sit ions – Quick Ref erence 01 2 3 4 5 6 7 8 9 A B C D E F Menu LEVEL 0 (0 0 to OF) Units PV Units Num. Selection PV Units Se con ds Se con ds % % mA mA Numeri- cal Numeri- cal PV .
IQ300IX.fm Page 5 Tuesday, October 2, 2001 1:43 PM.
*7ml19985GE01* Rev . 1.2 www .siemens-milltronics.com Siemens Milltronics Process Instruments Inc. 1 954 T ec hnology Dri ve, P .O. Box 4225 Peterborough, ON, Canada K9J 7B1 T el: (705) 745-2431 Fax: (705) 7 41 -0466 Email: techpubs@siemens-milltronics.
An important point after buying a device Siemens LC 500 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Siemens LC 500 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Siemens LC 500 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Siemens LC 500 you will learn all the available features of the product, as well as information on its operation. The information that you get Siemens LC 500 will certainly help you make a decision on the purchase.
If you already are a holder of Siemens LC 500, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Siemens LC 500.
However, one of the most important roles played by the user manual is to help in solving problems with Siemens LC 500. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Siemens LC 500 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center