Instruction/ maintenance manual of the product ID900 RTI
Go to page of 12
4075 East Mark et Street York, Pennsylvania 17402 Phone: 800-468-2321 or 717-840-0678 Fax: 717-755-8304 E-mail: rti@rtite ch.com Web-site: ww w .rtitech.co m OPERATION MANUAL ID900 Portable Refriger ant Identifier Manual No. 035-80979-00 WARNING Liquid Samples will Damage Unit Do not operate if filter has red flecks.
2 Table of Contents I. Cautions and Warnin gs 3 II. Functional Descript ion 4 III. Hardware Descrip tion 5 A . I n s t r u m e n t 5 1 . C a s e 5 2. Hanger 5 3. Sample Filter 5 4. Contr ol Panel 5 5. Air Intake Port 5 6. Sam ple Exhaust Por t 5 7. Sample Inlet Port 5 8.
3 I. Cautions and Warnings A. Cautions Never use any sam ple hose other than those appr ove d for use wit h the ID900. The use o f ot her hose t y pe s may introduc e errors and excessive refrigerant loss.
4 II. Functional Description Contam i nat i on of r efrigerant in ai r conditioni ng sy st em s or stor age cy linders can lead t o com ponent cor rosi o n, com ponent destruction, elevated system head pressure an d system failures when utiliz ed by unsu specting technicians.
5 III. Hardware Description A. Instrument : The instrument houses the b a sic com ponents requi r ed for operat i o n of t h e I D 900 as dep i ct ed abov e.
6 B. Sample Hoses 1. R12 Sam pl e H ose: i s a 6- foot hose consis ti ng of a ny l on inner tube surr oun ded by a po l y ur et hane out er tube. The hose contains the instrument connection fitting on one end and a ¼” SAE fe male flare nut on the other end.
7 IV. Operation Procedure A. Pre-Operati ve Inspection 1. Inspect the outside dia meter of the white element co ntained within the sample filter for an y signs of red spots. The appearance of red sp ots anywhere on the outside diameter of th e element ind icates that the filter req uires replacement to avoid foulin g o f t he i nst r um e nt .
8 5. When prom pted by the instrum ent, conne ct the free end of the samp le hose to the vehicle ai r conditioning system or refrigerant st orage cy l i nder por t and press the “A” pus hb utt on. Sample gas wil l imm ediately begin t o flow into the instrument.
9 V. Maintenance A. Error Codes and Corrective Actions The instrument is supp lied with self-diagn ostic software to aid the user in th e determination of proble ms. The software will supply error codes that will direct th e user towards specific corrective actions in the event o f system failure.
10 Warning: Should it be determined that th e air detectio n calibra tion failure has been caused by the presence of oxygen depleting compou nds i n t he surr ou ndi ng atmosphere t he use MU ST t a ke imm ediate action to ri d t he area of such compounds or provide adequate ventila tion to prevent such accumula tion.
11 c) When the hose is co mpletely dry insp ect the hose for signs of wear as described abo ve. 3. Air Detection Sensor Replacem ent Inspection Frequency: The i nstr um ent will inform the user of possible ai r det ect ion r eplacement r e qui rem ents with the error code Replace Air Sensor.
12 VI. Specifications Sample Parameters: Vapor only , oi l and l i qui d fr ee, 300 psig ( 2 M P a) M a xim um Detected Compounds: R12, R134a, Ai r and unk n ow n com pounds Sensor Technologies: Non-Dispersive Infrared and Chemical Fuel Cell Refrigerant Sample Size: 0.
An important point after buying a device RTI ID900 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought RTI ID900 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data RTI ID900 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, RTI ID900 you will learn all the available features of the product, as well as information on its operation. The information that you get RTI ID900 will certainly help you make a decision on the purchase.
If you already are a holder of RTI ID900, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime RTI ID900.
However, one of the most important roles played by the user manual is to help in solving problems with RTI ID900. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device RTI ID900 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center