Princeton Instruments 1 of 8 May 26, 2004 E:Ma nualsSyst emsPI-MAXPI-M AX2 DIF Ca mera.doc PI-MAX2 DIF Camera (Dual Image Feature) Introduction This technical note assumes a basic understanding of the operation of the PI-MAX2 and PTG in Gate Mode using WinView or Wi nSpec.
PI-MAX2 DIF Camera May 26, 2004 2 of 8 Princeton Instruments E:Ma nualsSyst emsPI-MAXPI-M AX2 DIF Ca mera.doc it is shifted under the masked area and held.
PI-MAX2 DIF Camera Princeton Instruments 3 of 8 May 26, 2004 E:Ma nualsSyst emsPI-MAXPI-M AX2 DIF Ca mera.doc Dual Trigger Mode The Dual Trigger Mode requires two triggers to acquire both images of the DIF operation. The PTG will control the gating of the intensif ier for both images.
PI-MAX2 DIF Camera May 26, 2004 4 of 8 Princeton Instruments E:Ma nualsSyst emsPI-MAXPI-M AX2 DIF Ca mera.doc Setup Hardware PTG ST -133 Contr oller Computer Spectrograph Acton 300i** GPIB Timing Gen Signal PI-MAX2 EXT . TRIG. IN PRE. TRIG. IN TIMING GEN.
PI-MAX2 DIF Camera Princeton Instruments 5 of 8 May 26, 2004 E:Ma nualsSyst emsPI-MAXPI-M AX2 DIF Ca mera.doc 3. The PI-MAX2 must be set to Gate mode for the intensifier to operate properly. Either click on the Gate mode button on the Custom Toolbar or select Gate mode on the Acquisition|Experiment Setup…|Main tab.
PI-MAX2 DIF Camera May 26, 2004 6 of 8 Princeton Instruments E:Ma nualsSyst emsPI-MAXPI-M AX2 DIF Ca mera.doc 4. Now the PTG needs to be programmed to m atch the experiment. In the Setup menu, select Pulser and click on Pulser Setup (or click the Pulser Setup button in the Custom Toolbar).
PI-MAX2 DIF Camera Princeton Instruments 7 of 8 May 26, 2004 E:Ma nualsSyst emsPI-MAXPI-M AX2 DIF Ca mera.doc the phosphor emission to drop to 10% of its peak value. The decay is not a simple single exponential; even after 100 µs there may be 1% or more of the first image on the phosphor screen.
PI-MAX2 DIF Camera May 26, 2004 8 of 8 Princeton Instruments E:Ma nualsSyst emsPI-MAXPI-M AX2 DIF Ca mera.doc This page intentionally left blank..
An important point after buying a device Princeton PI-MAX2 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Princeton PI-MAX2 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Princeton PI-MAX2 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Princeton PI-MAX2 you will learn all the available features of the product, as well as information on its operation. The information that you get Princeton PI-MAX2 will certainly help you make a decision on the purchase.
If you already are a holder of Princeton PI-MAX2, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Princeton PI-MAX2.
However, one of the most important roles played by the user manual is to help in solving problems with Princeton PI-MAX2. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Princeton PI-MAX2 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center