Instruction/ maintenance manual of the product M5 Omron Healthcare
Go to page of 12
Instruction Manual AUT OMA TIC BLOOD PRESSURE MONIT OR Model M5 ENGLISH (M5)-5 9/18/02 3:20 PM Page 1.
2 page Know Y our Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 Battery Installation / Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 Care and Maintenance .
3 Know Y our Unit Air Jack Air Plug AC Adapter Jack Systolic Blood Pressure Display Diastolic Blood Pressure /Pulse Display Arm Cuff Cuff T ube ST ART Button ON/OFF Button MEMORY Button 1. Slide the battery cover off in the direction of the arrow . 2.
4 Care and Maintenance T o protect your monitor from damage, please observe the following: Do not subject your monitor to extreme temperatures, humidity , or direct sunlight. Do not fold the cuff and tubing tightly . Do not inflate the monitor over 280 mmHg.
Systolic (mmHg) Diastolic (mmHg) Optimum < 120 < 80 Normal < 130 < 85 High-Normal 130 -139 85 - 89 Grade 1 Hypertension 140 -159 90 - 99 Grade 2 Hypertension 160 - 179 100 - 109 Grade 3 Hy.
1. A void eating, smoking, and exercising for at least 30 minutes before taking a measurement. Also rest for at least 15 minutes before taking your first reading. 2. Stress raises blood pressure. Avoid taking measurements during stressful times. 3. Remove tight-fitting clothing from your left upper arm.
7 How to Apply the Arm Cuff NOTE: If the circumference around your arm is greater than 32 cm you will need to use a large adult size cuff. The large adult cuf f is an accessory item and it is sold separately . Y ou may purchase this cuff by calling Omron representative or dealers.
8 How to T ake a Reading 1. Press the ON/OFF button. a) All display symbols appear for approximately one second. This is the initial LCD test. b) The display symbols disappear and the Deflation Indication symbol ( ) starts to flash.
9 How T o Use the Memory Function This monitor has a memory capable of storing fourteen (14) sets of readings. Every time you complete the measurement, the monitor automatically stores blood pressure and pulse rate. • Push the MEMORY button to recall stored readings while the Heart Symbol ( ) is displayed.
10 Error Symbol Cause Correction Cuff under-inflated. W ait 2-3 minutes and take another measurement. Repeat steps listed under “ How to T ake a Reading ” but manually inflate the cuff 40 mmHg higher than the No pulse or blood previous inflation value.
11 Specifications Model: M5 Display: LCD Digital Display Measurement Range: Pressure: 0-280 mmHg , Pulse: 40-180/minute Accuracy: Pressure: ±3 mmHg or 2% of the reading Pulse: ±5% of reading Inflati.
Product of OMRON Corporation, Japan OMRON HEAL THCARE SINGAPORE PTE L TD. 83 Clemenceau A venue #11-01, UE Square Singapor e 239920 ENGLISH (M5)-5 9/18/02 3:20 PM Page 12.
An important point after buying a device Omron Healthcare M5 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Omron Healthcare M5 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Omron Healthcare M5 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Omron Healthcare M5 you will learn all the available features of the product, as well as information on its operation. The information that you get Omron Healthcare M5 will certainly help you make a decision on the purchase.
If you already are a holder of Omron Healthcare M5, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Omron Healthcare M5.
However, one of the most important roles played by the user manual is to help in solving problems with Omron Healthcare M5. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Omron Healthcare M5 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center