http://www.omega.com e-mail: info@omega.com U ser ’ s Guide V R 2 0 0 / R S - 4 2 2 A C o m m u n i c a t i o n s I n s t r u c t i o n M a n u a l.
Servicing North America: USA: One Omega Drive, Box 4047 ISO 9001 Certified Stamford, CT 06907-0047 T el: (203) 359-1660 F AX: (203) 359-7700 e-mail: info@omega.com Canada: 976 Bergar Laval (Quebec) H7L 5A1 T el: (514) 856-6928 F AX: (514) 856-6886 e-mail: info@omega.
1 PREF ACE This instruction manual describes the optional RS-422-A communication function for the VR200 view recorder. For details concerning the operations of the recorder, refer to the instruction manual for the VR200. Trademarks • “IBM” is a trademark of International Business Machines Corporation.
2 CONTENTS PREFACE Chapter 1 INSTALLING THE RS-422-A INTERFACE 1.1 Interface Specifications .................................................................................................... ............ 1-1 1.2 Interface Terminal ..................
3 CONTENTS Chapter 1 Chapter 2 Chapter 3 Chapter 4 Chapter 5 Chapter 6 Chapter 7 INDEX APPENDIX 2.4 Escape Sequence ............................................................................................................. ............ 2-13 2.4.1 Execution of Trigger .
1 - 1 1 INST ALLING THE RS-422-A INTERF ACE 1.1 Interface Specifications Chapter 1 INST ALLING THE RS-422-A INTERF ACE The option /C3 provides the VR200 view recorder with the EIA (Electronic Industri.
1 - 2 1.2 Interface Terminal 1.2.1 Terminal Assignment MEMORY SD A SD B SG G (signal ground) RD A RD B FG G (frame ground) Connect RD to TD of the host computer. Connect SD to RD of the host computer. Option terminal Indication of RS-422-A terminals Figure 1.
1 - 3 1 INST ALLING THE RS-422-A INTERF ACE 1.3 Communication Wiring If the host PC is equipped with an RS-422-A interface, the VR200 can be connected directly. If the host PC is equipped with only an RS-232-C interface, the VR200 can be connected using a converter (SHARP model Z-101HE or equivalent).
1 - 4 If A < B, then signal = 1 If A > B, then signal = 0 1.4.1 Start-stop (Asynchronous) Communication The built-in RS-422-A interface communicates based on the start-stop (asynchronous) system.
1 - 5 1 INST ALLING THE RS-422-A INTERF ACE Note The VR200 identifies an LF as the end of a text when receiving data (and sends a CR and LF as the terminator when sending data). As shown in the example in Figure 1.6, when a CR and LF are used as the terminator, the CR is ignored.
1 - 6 1.5 Setting up the RS-422-A Communication Interface SETUP PROCEDURE: 1 Unpack the VR200 recorder. 2 While pressing the [ ] key, turn on the power to enter the SETUP mode. 3 Use the [ ▲ ] / [ ▼ ] keys to select the display ‘SETUP=OPT,’ then press the [ ] key.
2 - 1 2 RECEIVING FUNCTIONS Chapter 2 RECEIVING FUNCTIONS This chapter describes the program set commands and program control commands. Remember first to open a device by invoking the ESC O command before the set or control commands can be sent.
2 - 2 2.2 Program Set Commands 2.2.1 List of Program Set Commands Command Type Command Function Set SR SA SN SW SD SY SZ SP SK ST SL SF SG SC SM SH SX MD Sets the input range and display scale. Sets the alarms. Sets the engineering unit. Sets the waveform span rate.
2 - 3 2 RECEIVING FUNCTIONS Input range table Input type Range Input Measurement range Unit DC Voltage 20 mV 20 mV -20.00 to 20.00 mV mV 60 mV 60 mV -60.00 to 60.00 mV mV 200 mV 200 mV -200.0 to 200.0 mV mV 2 V 2 V -2.000 to 2.000 V V 6 V 6 V -6.000 to 6.
2 - 4 Example: SR01, SCL, VOLT, 20mV, 0, 1000, –1000, 1000, 1 This example performs 0 to 10 mV input in channel 01 and is scaled from – 100.0 to 100.0. Note An error will occur if one or two of p5, p6, and p7 are omitted. However, it is possible to omit all three values if you do not want to change them.
2 - 5 2 RECEIVING FUNCTIONS 2.2.4 Setting the Unit (See Section 4.3 of the instruction manual for the VR200.) Format: SNp1, p2 p1: Channel number (CC) p2: Engineering unit (up to six characters) Example: SN02, kg This example assigns the unit “ kg ” to channel 2.
2 - 6 2.2.8 Setting the Discrete (ZONE) Display (See Section 4.7.1 of the instruction manual for the VR200.) Format: SZp1, p2, p3 p1: Channel number (CC) p2: Lower display boundary value (0 to 95) p3:.
2 - 7 2 RECEIVING FUNCTIONS 2.2.12 Setting the Trip Level (See Section 4.7.5 of the instruction manual for the VR200.) Format: SLp1, p2, p3 p1: Trip level number VR100 and VR200 with style number 1 or lower 1 for level 1 (red line) 2 for level 2 (yellow line) VR200 with style number 2 higher 1 to 6 p2: Whether to display.
2 - 8 2.2.15 Setting the LCD Brightness and LCD Saver (LCD) (See Section 4.8 of the instruction manual for the VR200.) Format: SCp1, p2, p3 p1: LCD brightness level. 0 to 15 p2: Sets the LCD saver function “ ON ” or “ OFF. ” p3: Timer setting to activate the LCD saver.
2 - 9 2 RECEIVING FUNCTIONS If your model has -3M option, add the parameter p3. p3: Select the methods when saving measurement data to the floppy disk: ALL (default) save all measurement data from the VR200 internal memory.
2 - 10 2.3 Program Control Commands 2.3.1 List of Program Control Commands Command Type Command Function Control UD AK MI EV BO TS FM LF LO LI ME UM Returns the screen to the standard screen. Acknowledges the alarm(s). Initializes the data memory. Writes a time-axis mark or raises a manual trigger.
2 - 11 2 RECEIVING FUNCTIONS 2.3.7 Selecting the Output Data Command Function TS0 Requests the measured values. TS1 Requests the parameter set values. TS2 Requests unit and decimal point information.
2 - 12 2.3.12 Message Writing(for the VR 200 with style number 2 or higher) Command Function MEp1 Selects message number for which parameter set values to be displayed and the information file is saved.
2 - 13 2 RECEIVING FUNCTIONS 2.4 Escape Sequence 2.4 Escape Sequence The following escape sequence commands are used to control the communications. 2.4.
2 - 14 Note A syntax error status is set when detected and reset when read by the host PC. On the contrary, the Memory End status still reflects the current status even after being read.
3 - 1 3 TRANSMITTING FUNCTIONS 3.1 Introduction to Output Data Formats Chapter 3 TRANSMITTING FUNCTIONS This chapter describes the different output formats. 3.1 Introduction to Output Data Formats The type of data to be sent from the VR200 recorder is specified by the TS command (TS0, TS1, or TS2: see Section 2.
3 - 2 3.2 Output Data Formats There are four formats in which data can be output. • TS0 + ESC T + FM0 (outputs measured values in ASCII mode) • TS0 + ESC T + FM1 (outputs measured values in Binary mode) • TS1 + ESC T + LF (outputs parameter setting values) • TS2 + ESC T + LF (outputs information on engineering unit and decimal point) 3.
3 - 3 3 TRANSMITTING FUNCTIONS 3.2.2 Output Format of Measured Values in Binary Mode When the TS0, ESC T and FM1 commands are received, the measured values and computed results are output in binary.
3 - 4 (4) Measured data 12345 Byte A 2 A 1 A 4 A 3 Alarm status (for each level) A1 to An correspond to statuses for levels 1 to 4. Alarm status Description 1 2 3 4 5 6 0 No alarm or the alarm functio.
3 - 5 3 TRANSMITTING FUNCTIONS 3.2.4 Output Format of Unit and Decimal Point Information When receiving the TS2, ESC T, and LF commands, the VR200 recorder sends the information on the unit and decimal point in the following format. The channel numbers can be specified with the LF command.
4 - 1 4 TIME CHAR T Chapter 4 TIME CHAR T One sampling (125 ms) When outputting binary data * A/D and primary computation ESC T command from host Output of data * If the ESC T command is received to output data before the data is updated, the previous sam p le data will be out p ut.
5 - 1 5 INITIAL ST A TUS Chapter 5 INITIAL ST A TUS The initial status after power on is as follows. TS0 Output data type is the measured values. FM0, 01, 04 Measured values are output as ASCII codes. From channel 01 To channel 04 (All channels are specified.
6 - 1 6 ERRORS DURING RS-422-A OUTPUT 6.1 Preventing Errors Chapter 6 ERRORS DURING RS-422- A OUTPUT 6.1 Preventing Errors Do not send an FM or LF command from the host PC until the measured data or set point data in all the channels specified by the last command have been output.
6 - 2 6.2 How to Request Error Message Output If an error occurs when the host sends a setting or control command to a VR200 recorder via the RS-422-A communication interface, the host can request the VR200 to output an error message.
6 - 3 6 ERRORS DURING RS-422-A OUTPUT 6.3 Timing of Resetting the Error Status 6.3 Timing of Resetting the Error Status When the VR200 receives an ESC S command after an error occurred, the recorder outputs the corresponding error message and the error status is reset at the same time.
7 - 1 7 PROGRAM EXAMPLES Chapter 7 PROGRAM EXAMPLES Examples of application programs for an IBM PC are shown below. Environment Computer used: IBM PC Mode: Character length = 8 bits, parity = none, stop bit = 1, baud rate = 1200 bps Handshake: None The file name used for writing to and reading from the floppy disk is TEST.
7 - 2 7.2 Program to Read Parameter Settings from Disk, Display Them on Screen, and Set up VR200 100 ‘************************************************************ 110 ‘** RS422A Example Program fo.
7 - 3 7 PROGRAM EXAMPLES 7.3 Program to Read Unit and Decimal Point Information from VR200, Display It on Screen, and W rite It to Disk 100 ‘*********************************************************.
7 - 4 7.4 Program to Output Measured Data (in ASCII Code) from VR200 and W rite to Disk 100 ‘************************************************************ 110 ‘** RS422A Example Program for IBM PC .
7 - 5 7 PROGRAM EXAMPLES 7.5 Program to Output Measured Data (in Binary Code) from VR200 and W rite to Disk 100 ‘************************************************************ 110 ‘** RS422A Example.
App. - 1 IM 4N1A1-1 1E APPENDIX Appendix 1 ASCII Code Table APPENDIX Appendix 1 ASCII Code T able Note • Select the degree symbol ( ° ) of ° C or ° F as follows: (a) In case of measured values ou.
App. - 2 IM 4N1A1-1 1E Appendix 2 File Naming Rules File Name Length Data file: Up to seven alphanumeric characters Configuration (parameter settings) file: Up to eight alphanumeric characters Lowercase/Uppercase Letters All lowercase letters are processed as uppercase letters for file names, based on the rules of MS-DOS.
Index - 1 INDEX INDEX A alarms acknowledgment ............................................. 2-10 setting ................................................................ 2-4 ASCII codes ....................................................... App-1 C channel selection .
Index - 2 range settings DELT ........................................................... 2-2 DI (Digital Input) ........................................ 2-3 RTD ............................................................. 2-2 SCL (Scaling) .........
W ARRANT Y / DISCLAIMER OME GA EN GINE ERIN G, IN C. wa rrant s th is unit to b e free o f def ects i n m a terial s an d wor kma nshi p for a pe riod of 2 5 m ont hs f r o m d ate of purch ase .
M3285/0299 W h e r e D o I F i n d E v e ry t h i n g I N e e d f o r P r o c e s s M e a s u r e m e n t a n d C o n t r o l ? O ME G A … O f C o u r s e ! T E M P E R AT U R E Th erm ocoupl e , R .
An important point after buying a device Omega Speaker Systems VR200 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Omega Speaker Systems VR200 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Omega Speaker Systems VR200 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Omega Speaker Systems VR200 you will learn all the available features of the product, as well as information on its operation. The information that you get Omega Speaker Systems VR200 will certainly help you make a decision on the purchase.
If you already are a holder of Omega Speaker Systems VR200, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Omega Speaker Systems VR200.
However, one of the most important roles played by the user manual is to help in solving problems with Omega Speaker Systems VR200. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Omega Speaker Systems VR200 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center