USING THE RPS-1in the OM-420 Omega Engineering Inc. 1 Stamford, CT USING THE RPS-1, RECHARGEABLE POWER SUPPLY IN THE OM-420 OVERVIEW: The RPS-1 is a self-contained battery power source for use in conjunction with the OM-320/420 System Base to power transducers requiring external power (e.
USING THE RPS-1in the OM-420 Omega Engineering Inc. 2 Stamford, CT An integral State-of-Charge (SOC) indicator is provided which will display an approximate Battery State-of- Charge based on battery voltage under a fixed load. Test is actuated by depressing a momentary SOC Test Switch.
USING THE RPS-1in the OM-420 Omega Engineering Inc. 3 Stamford, CT CONTROL INPUT Two terminal connections are provided for connection of a low voltage power supply control signal. This input signal controls the ON/OFF cycling of power supplies set to the AUTO mode.
USING THE RPS-1in the OM-420 Omega Engineering Inc. 4 Stamford, CT The MODE switch (bottom or eighth switch) should be set for the desired operation, AUTO or ON. AUTO mode configures the supply to cycle ON and OFF per the state of the control signal coming from an external device (OM-320/420).
USING THE RPS-1in the OM-420 Omega Engineering Inc. 5 Stamford, CT voltage is set to 12 VDC or higher output. For 12VDC or higher power supply OUTPUT VOLTAGE settings, the MAIN POWER SWITCH must be set to 24VDC. In the 12VDC mode, the two batteries are in parallel, hence the amp hour capacity is doubled (i.
USING THE RPS-1in the OM-420 Omega Engineering Inc. 6 Stamford, CT held down for approximately 10 seconds to dissipate any battery surface charge and give the most consistent reading. Best SOC readings are given with both power supplies off and no charging current.
USING THE RPS-1in the OM-420 Omega Engineering Inc. 7 Stamford, CT BATTERY REPLACEMENT NOTE: USE CAUTION IN WORKING WITH THE GEL-CELL BATTERIES TO PREVENT SHORTING OF THE TERMINALS OR WIRING AT ALL TIMES! Under reasonable operating conditions and temperatures, the two gel-cell batteries should provide several hundred charge/discharge cycles.
USING THE RPS-1in the OM-420 Omega Engineering Inc. 8 Stamford, CT PROBLEM: Batteries are fully discharged. SOLUTION: Charge the batteries. SYMPTOM: STATE-OF-CHARGE INDICATOR SHOWS LESS THAN FULL PROB.
An important point after buying a device Omega Engineering OM-420 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Omega Engineering OM-420 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Omega Engineering OM-420 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Omega Engineering OM-420 you will learn all the available features of the product, as well as information on its operation. The information that you get Omega Engineering OM-420 will certainly help you make a decision on the purchase.
If you already are a holder of Omega Engineering OM-420, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Omega Engineering OM-420.
However, one of the most important roles played by the user manual is to help in solving problems with Omega Engineering OM-420. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Omega Engineering OM-420 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center