Instruction/ maintenance manual of the product Dissolved Oxygen System Omega Engineering
Go to page of 29
omega.com e-mail: info@omega.com For latest product manuals: omegamanual.info DOCN600 Dissolved Oxygen System Shop online at U ser ’ s Guide.
Servicing Nor th America: U.S.A.: One Omega Drive, P.O. Box 4047 ISO 9001 Certified Stamford, CT 06907-0047 TEL: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com Canada: 976 Bergar Laval (Quebec) H7L 5A1, Canada TEL: (514) 856-6928 FAX: (514) 856-6886 e-mail: info@omega.
TABLE OF CONTENTS GENERAL INFORMATION 1 SPECIFICATIONS 2 INSTALLATION 4 DESCRIPTION OF FUNCTIONS 8 START-UP AND OPERATION 13 OPERATING HINTS 17 UTILITY MENU 18 DIAGNOSTICS 20 TROUBLESHOOTING AND SERVI.
Page 1 MODEL DOCN600 DISSOLVE D OXYGEN ANALYZERS AND MODEL DOE-601 DISSOL VED OXYGEN SENSOR INSTRUCTION MANUAL 1 . 0 G E N E R A L I N F O R M A T I O N The DOCN600 Dissolved oxygen analyzer is a versati le indu strial microprocessor based instrument.
Page 2 2.0 SPECIFICATIONS DISPLAY: 3 1/2 digit LED, 1/2" high digits MEASUR ING RANGES: D.O.: 0-20.00 ppm and 0-100% s aturation, switch selectable Temperature: 0-40°C (32 to 104 °F) POWER REQUIREMENTS : 98-132 Vac, 50/60 Hz (less than 10 VA) Optional: 196-264 Vac, 50/60 Hz AMBIEN T CONDITIONS: -30 to 50°C (-22 to 122°F) 0 to 90% R.
Page 3 MODEL DOCN600 ADDITIONAL FEATURES CONTRO L RELAYS: Two relays can be independently set for operation in response to rising or falling value and for fail-safe operation. Deadband (hysteresis) independently adjustable. Rating: 5A 115/230 Vac, 5A 30 Vdc.
Page 4 3.0 INSTALLATION 3 . 1 L o c a t i o n 3.1.1 Locate the instrument within the re ach of the cable p rovided for the DOE-601 s e n s o r . 3.1.2 Select an installation site which is: • free of.
Page 5 3.5 Sensor Installation, Flotation Mounting 3.5.1 Refer to diagram below and proceed as follows: a) Fasten elbow to 90-inch long pipe NOTE : Thread seala nt (Teflon tape) is recommended on all .
Page 6 d) The sensor may be attached to float assembly in one of two ways: 1. With sensor EXTENDED FROM BOTTOM of flotation ball: Insert 12-inch long extender pipe into th e flotation ball and faste n it to the end of adapter pipe. This will extend the senso r ’ s membrane 1 2 inches below the flotation ball.
Page 7 h) Install pivot mounting bracket and moun t junction box on the bracket. Insert pipe/sensor assembly through the bracket to the desired position. Tighten pi pe locking screw. 3.6 Sensor Installation, Submersion Mounting 3.6.1 a) Refer to diagram below and proceed as follows: b) Route sensor cable throu gh pipe and fasten sensor to pipe.
Page 8 4.0 DESCRIPTION OF FUNCTIONS 4 . 1 O v e r v i e w 4.1.1 The DOCN600 is a microprocessor ba sed dissolved oxygen analyzers are designed for industrial applications. They operate in conjunction with the DOE-601 sensor. The software in the unit makes the instru ment very easy to operate and maintain.
Page 9 Some of these are: • Recall and easy adjustment of relay and output pa rameters • Push button calibration • A HOLD function for outputs • Continuous sensor chec k during measurement •.
Page 10 4.6 Analog Outputs 4.6.1 The analog output signals con s i s t of a non-i s ol ated 0-1 mA, 0-5 Vdc, and isolated 4- 20 mA signals. 4.6.2 From the factory all of the analog outputs have a linear range corresponding to the full range of the instrument.
Page 11 4.7.4 When in the menu mode, the display initially shows the current value of the parameters, such as the Setpoint of the c ontrol relay, while putting all of the outputs on hold. The two arrow buttons are u s ed to adjust the display value up or down.
Page 12 Dip Switch Bank S2 1 Fail-safe for Relay B NO YES 2 Direction of Control for Relay B Rising Falling 3 Reserved 4 Reserved 5 Reserved 6 Reserved 7 Reserved 8 Reserved 4 .
Page 13 5.0 START-UP AND OPER ATION 5 . 1 P a s s w o r d 5.1.1 To enter the menu press CALL then PASSWORD will be indicat ed. With each press of CALL you will step through the m enu. When the last item is rea c h ed the menu wraps around to “%sat/ppm ”.
Page 14 i) If you wish to change the measuring unit to ppm, pre ss CALL a number of times to return to “%sat/ppm” and change pp m with the down button. j) Press RUN to return to on line measurement. NOTE : If a major change is n ecessary to make the display read 100% the display will flash when you press ENTER.
Page 15 5.6 Temperature 5.6.1 The temperature of the process can be read at any time by entering the menu and calling for TEMPERATURE. Either °C or °F will be indicated dep ending on the position of DIP switch No.
Page 16 5.8.2 The relay setpoint may be at any place on the scale. To establi sh the setpoint proceed as follows: a) Press CALL to enter the MENU. If the PASSWORD function is enabled, enter the password then press CALL to illuminate the “%sat /ppm” indicator.
Page 17 5 . 1 2 A l a r m H i g h 5.12.1 The instrument is fitted with a relay, which is set to activate on both high and low alarm conditions. The deadban d is factor y set. To set the ALARM HIGH proceed a s f o l l o w s . a) Enter the menu and press CALL until AL ARM HIG H is indicate d.
Page 18 6 . 3 R e s e t 6.3.1 The instrument can be reset without losing calibration or any of the stored values by pressing and rele asing the S9 reset button (located at the bottom of the board near the center) This action is equivalent to turning the po wer off and on.
Page 19 7.3 Temperature Output 7.3.1 The 0-5 Vdc and 0-1 mA outputs may be pro grammed to track the temperature of the process. The factory outputs have a li near range co rresponding to 0°C to 40°C, (or 32°F to 104°F.) Suppose you wish the output to span 10°C to 45 °C.
Page 20 b) Turn on the power. Enter the Utility Menu as described in 7.2.1. c) Press CALL to indicate CALIBRATION, which is "4 -20 mA Output adjust, low" in the Utility Menu (See table in 7.2.2.) No w use the arrow keys to make you r DVM read 8 mA.
Page 21 9.0 TROUBLESHOOTING AND S ERVICE 9.1 Isolate the cause 9.1.1 When a measurement problem occurs, the fi rst step is to try to isolate the cause. If the DOCN600 is powered, go through the menu and check your settings. A convenient way to do this is to call TEST.
Page 22 e) Now proceed to turn each of the other S43 switches On with the remaining switches Off. The display should read within 5% of 25°C with S witch 2 On and 35°C with Switch 3 On. If this is the case the analyzer is in order and the problem is in the sensor.
Page 23 d) After all the above op erations are performed the FAIL LED will turn off and STATUS LED will turn green and Status “0” should be obtaine d. e) The unit may be tested now, using the TEST menu item (See Section 5.5), or the built-in self-testi ng feature, (See Section 9.
Page 24 STATEMENTS OF CONFORMITY FROM THE MANUFACTURER U.S.A. WARNING: This equipment gen erates, uses, and can radiate radio frequenc y energy and if not installed and used in accordance w ith th e instructions manual, may cause interference to radio communications.
W ARRANTY / DISCLAIMER OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months fr om d ate of pu rcha se. OMEGA ’ s W ARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time.
M4177/0 8 05 Where Do I Find Ever ything I Need for Pr ocess Measurement and Contr ol? OM EGA…Of Course! Shop online at omega.com TEMPERA TURE MU Thermocouple, RTD & Thermistor Probes, Connector.
An important point after buying a device Omega Engineering Dissolved Oxygen System (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Omega Engineering Dissolved Oxygen System yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Omega Engineering Dissolved Oxygen System - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Omega Engineering Dissolved Oxygen System you will learn all the available features of the product, as well as information on its operation. The information that you get Omega Engineering Dissolved Oxygen System will certainly help you make a decision on the purchase.
If you already are a holder of Omega Engineering Dissolved Oxygen System, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Omega Engineering Dissolved Oxygen System.
However, one of the most important roles played by the user manual is to help in solving problems with Omega Engineering Dissolved Oxygen System. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Omega Engineering Dissolved Oxygen System along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center