CDE-440 Series Conductivity Process Cells.
.
Instr uctions Manual Conductivity Cells CDE-440-001/CDE-440-01/CDE-440-1/CDE-440-5/CDE-440-01T/CDE-442 Inde x C o n d u c t i v i t y C e l l s .............................................................3 M e c h a n i c a l D e s c r i p t i o n ..
1. Certificado de Garantia 3. Conductivity Cells 2 Cell is composed basically by two boards or two concentric cylinders(electrodes), with well defined geometry . The sp ace between the boards or cylindres creates a solution collumn, being so the measurement is independent from the total solution volume.
Insertion Cell CDE-440-01 Cell Models CDE-440-001, CDE-440-01, CDE-440-1, CDE-440-5 and CDE-440- 01T are called Insertion or Immersion Cells. 1. Certificado de Garantia 4. Mechanical Description Cell Model CDE-442 is called T oroydal or Electrodeless Cell, as it does not offer electrodes.
1. Certificado de Garantia 5. T echnical S pecifications Model CDE - 440 - 001 CDE - 440 - 01 CDE - 440 - 1 CDE - 440 - 5 CDE - 440 - 1T CDE - 442 Ty p e Inserti on Torroid Range (S/cm ) 0.01µS to 2mS 0.1µS t o 20m S 1µS to 100mS 10µS to 150 mS 0.
Green Y ellow Red Black Cable Knockout Connection Box Connection Box Interconnection Base Interconnection Base 3 4 1 2 1. Certificado de Garantia 6. Inst allation 5 CN1 CN1 1 1 CN3 CN3 CN2 CN2 CN4 CN4 + + RS485 RS485 S2 S2 mA-2 mA-2 2 2 3 3 S3 S3 - - + + - - mA-1 mA-1 AC AC - - + + TC-401 TC-401 Na Na S1 S1 C C Vm.
a) Do not cut on mend the cell cables! Interconnection Cable This cell type is also connected directly to the equipment connection barr, so cable terminal pin type are used for some equipment cases. Colorless from Red Red Black Colorless from Shield Blue Colorless from Blue 1.
T orroydal cell CDE-442 generally must be installed in a T ransfluency Body made out of lower magnetic permeability material. S pecial attention must be observed regarding the sample flow being Flanged Threaded Point 1. Certificado de Garantia 6. Inst allation (cont.
1. Certificado de Garantia 7. Maintenance Cleaning Calibration Calibration for L ow Conductivity R eading(Ultra Pure W ater) 8 A ll OMEGA equipments, are shipped from factory tested and calibrated.
1. Certificado de Garantia 8. T roubleshooting 9 1) When Calibrating the equipment, the function is not complete V erify for humidity presence at probes header or at connection box of the equipment. Another possibility could be the equipmen ts interconnection cable and cell being cut or in short.
.
M-4506/0908.
An important point after buying a device Omega CDE-440 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Omega CDE-440 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Omega CDE-440 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Omega CDE-440 you will learn all the available features of the product, as well as information on its operation. The information that you get Omega CDE-440 will certainly help you make a decision on the purchase.
If you already are a holder of Omega CDE-440, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Omega CDE-440.
However, one of the most important roles played by the user manual is to help in solving problems with Omega CDE-440. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Omega CDE-440 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center