Instruction/ maintenance manual of the product IRC5PO2W Invacare
Go to page of 32
Dealer: This manual MUST be given to the end user of this product. User: BEFORE using this product, read this manual and save for future reference. User Manual For more information regarding Invacare products, parts, and services, please visit www.invacare.
WARNING Perfecto 2 ™ Series 2 Part No 114348 2 WARNING DO NOT USE THIS PRODUCT OR ANY AVAILABLE OPTIONAL EQUIPMENT WITHOUT FIRST COMPLETELY READING AND UN DERSTANDING THESE INSTRUCTIONS AND ANY ADDITIONAL INSTRUCTIONAL MATERIAL SUCH AS OWNER’S MANUALS, SERVICE MANUALS OR INSTRUCTION SHEETS SUPPLIED WITH THIS PRODUCT OR OPTIONAL EQUIPMENT.
TABLE OF CONTENTS Part No 1143482 3 Perfecto 2 ™ Series Table of Contents ACCESSORIES .................................................. 2 SPECIAL NOTES ............................................... 4 LABEL LOCATION ...............................
SPECIAL NOTES Perfecto 2 ™ Series 4 Part No 114348 2 SPECIAL NOTES Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage.
SPECIAL NOTES Part No 1143482 5 Perfecto 2 ™ Series CAUTION “Caution: Federal law restricts this device to sale or rental by or on order of .
LABEL LOCATION Perfecto 2 ™ Series 6 Part No 114348 2 LABEL LOCATION FOR OPERA TING INSTRUCTIONS or ALARMS (See Operator’s Manual) DANGER - ELECTRIC SHOCK DANGER - RISK OF FIRE NO SMOKING - Keep ALL sources of ignition out of the room in which this product is located and away from areas where o xygen is being delivered.
SECTION 1—GENERAL GUIDELINES Part No 1143482 7 Perfecto 2 ™ Series SECTION 1—GENERAL GUIDELINES In order to ensure the safe installation, assembly and operation of the Pe r f e c t o 2 concentrator these instructions MUST be followed.
SECTION 1—GENERAL GUIDELINES Perfecto 2 ™ Series 8 Part No 114348 2 Operating Information For optimum performance, Inv acare recommends that each concentrator be on and runni ng for a minimum of 30 minutes at a ti me.
SECTION 1—GENERAL GUIDELINES Part No 1143482 9 Perfecto 2 ™ Series Other devices may experience interference from ev en the low levels of electromagnetic emissions permitted by the above standards.
SECTION 2—FEATURES Perfecto 2 ™ Series 10 Part No 1143482 SECTION 2—FEATURES Oxygen Outlet Flowmeter Oxygen Purity Indicator Lights / Fault an d Power Indicator Lights Power Switch Circuit Break.
SECTION 3—HANDLING Part No 1143482 11 Perfecto 2 ™ Series SECTION 3—HANDLING The concentrator should AL W A YS be kept in the upright position to prevent cabinet damage while being transported.
SECTION 4—TYPICAL PRODUCT PARAMETERS Perfecto 2 ™ Series 12 Part No 1143482 SECTION 4—TYPICAL PRODUCT PARAMETERS Electrical Requirements: 120 VAC ± 10% (132 VAC/10 8 VAC), 60 Hz Rated Current I.
SECTION 4—TYPICAL PRODUCT PARAMETERS Part No 1143482 13 Perfecto 2 ™ Series Change in maximum recommended flow when back pressure of 7kPa is applied: 0.7 L/min Filters: Cabinet, Outlet HEPA and Compressor Inlet Safety System: Current overload or line surge shutdown.
SECTION 4—TYPICAL PRODUCT PARAMETERS Perfecto 2 ™ Series 14 Part No 1143482 *NOTE: Stated concentration levels achieved after initial warm ‐ up period (appr oximately 30 minutes). Placement: No closer than three inche s from any wall , furniture, draperies, or similar surfaces.
SECTION 5—OPERATING INSTRUCTIONS Part No 1143482 15 Perfecto 2 ™ Series SECTION 5— OPERATING INSTRUCTIONS Introduction Yo u r oxygen concentrator is intended for individual use indoors. It is an electronically operated device that separates oxygen from room air .
SECTION 5—OPERATING INSTRUCTIONS Perfecto 2 ™ Series 16 Part No 1143482 Set Up Plug in Power Cord Plug in pow er cord to an electrical outlet. Connect Humidifier (If So Prescribed) WARNING DO NOT overfill humidifier .
SECTION 5—OPERATING INSTRUCTIONS Part No 1143482 17 Perfecto 2 ™ Series FIGURE 5.2 Humidifer Bottle Adapter 5. Attach the humidifier bottle adapter to the humidifier .
SECTION 5—OPERATING INSTRUCTIONS Perfecto 2 ™ Series 18 Part No 1143482 FIGURE 5.3 Humidifier Compartmen t Power Switch NOTE: For this pr ocedure, refe r to FIGURE 5.4. 1. Press pow er sw itch to on position.
SECTION 5—OPERATING INSTRUCTIONS Part No 1143482 19 Perfecto 2 ™ Series Flowrate NOTE: For this procedure, refe r to FIGURE 5.5 on page 20. 1. T urn the flowrate knob to the setting prescribed by you r physician or therapist.
SECTION 5—OPERATING INSTRUCTIONS Perfecto 2 ™ Series 20 Part No 1143482 FIGURE 5.5 Flowrate Front Panel Indicators - Units with SensO 2 The SensO 2 feature monitors the purity of oxygen generated by the oxygen concentrator .
SECTION 5—OPERATING INSTRUCTIONS Part No 1143482 21 Perfecto 2 ™ Series Initial Startup of the Concentrator NOTE: Concentrator may be used during the initial sta rt warm ‐ up time (approximately 30 min.
SECTION 5—OPERATING INSTRUCTIONS Perfecto 2 ™ Series 22 Part No 1143482 Initial Startup of the Concentrator NOTE: Concentrator may be used during the initial sta rt warm ‐ up time (approximately 30 min).
SECTION 6—MAINTENANCE Part No 1143482 23 Perfecto 2 ™ Series SECTION 6—MAINTENANCE WARNING The Invacare concentrators are specifically designed to minimize routine preventiv e mainte nance.
SECTION 6—MAINTENANCE Perfecto 2 ™ Series 24 Part No 1143482 2. Clean the cabinet filter with a va c u u m cleaner or wa s h in w arm soapy wa t e r and rinse thoroughly . 3. Dry the filter thoroughly before reinstallation.
SECTION 6—MAINTENANCE Part No 1143482 25 Perfecto 2 ™ Series FIGURE 6.2 Preventi ve Maintenance Record ON EACH INSPECTION MODEL NO. I R C SERIAL NO. PREVENTIVE MAINTENANCE RECORD Record Date of Service Record Elapsed Hours On Hour Meter Clean Cabinet Filter(s) Check Prescribed L/min.
SECTION 7—TROUBLESHOOTING GUIDE Perfecto 2 ™ Series 26 Part No 1143482 SECTION 7—TROUBLESHOOTING GUIDE SYMPTOM: PROBABLE CAUSE: SOLUTION: Alarm: Short beeps, long pause Concentrator not operating, power switch On. Beep......Beep. ..... Main Power Loss: 1.
SECTION 7—TROUBLESHOOTING GUIDE Part No 1143482 27 Perfecto 2 ™ Series YELLOW or RED light Illuminated. Alarm: Continuo us On RED light only. * Only applies to IRC5PO2. 1. Low oxygen purity.* 1. Clean or Replace filters. 2. Kinked or blocked tubing, cannula or humidifier.
NOTES Perfecto 2 ™ Series 28 Part No 1143482 NOTES.
NOTES Part No 1143482 29 Perfecto 2 ™ Series NOTES.
NOTES Perfecto 2 ™ Series 30 Part No 1143482 NOTES.
LIMITED WARRANTY Part No 1143482 31 Perfecto 2 ™ Series LIMITED WARRANTY NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURE D AFTER JULY 4, 197 5. This warranty is extended only to th e original purchaser who purchases this product when new and unused from Invacare Corporation or a dealer.
LIMITATIONS AND EXCLUSIONS: THE FOREGOING WARRANTY SHALL NOT APPLY TO PRODUCTS SUBJECTED TO NEGLIGENCE, ACCIDENT, IMPROPER OPERATION, MAINTENANCE OR STORAGE, SOOT OR SMOKE-FILLED ENVI RONMENTS, OR OTH.
An important point after buying a device Invacare IRC5PO2W (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Invacare IRC5PO2W yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Invacare IRC5PO2W - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Invacare IRC5PO2W you will learn all the available features of the product, as well as information on its operation. The information that you get Invacare IRC5PO2W will certainly help you make a decision on the purchase.
If you already are a holder of Invacare IRC5PO2W, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Invacare IRC5PO2W.
However, one of the most important roles played by the user manual is to help in solving problems with Invacare IRC5PO2W. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Invacare IRC5PO2W along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center