Instruction/ maintenance manual of the product 4909 CPVC Honeywell
Go to page of 34
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 70-82-25-19 Rev. 1 7/99.
ii 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 Copyright, Notices, and Trademarks Printed in U.S.A. – © Copyright 1999 by Honeywell Inc.
7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual iii A bout This Document Abstract The purpose of this document is to support the installation operation and m aintenance of the 4909 CPVC Conductivity Cell Remov al/Insertion Assem bly.
iv 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 Organization Telephone Address Honeywell TAC 1-800-243-9883 Voice 1100 Virginia Drive Fort Washington, PA 19034.
7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual v Contents 1. INTRODUCTION ....................................................................................................1 1.1 Overview ............................
vi 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 4. MA INTENA NCE A ND REPLAC EMENT PA RTS .................................................17 4.1 To Clean The Cell ...................................................
7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual vii Tables Table 4-1 Replacem ent Parts __________________________________________________________ 18 Table 4-2 Voltage.
.
Introduction 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 1 1. Introduction 1.1 Overview The 4909 Conduct ivit y C ell Ins ertion/R emoval A ssembly is design ed fo.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 2 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 1.
Introduction 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 3 1/ 4" NP T Plug Cabl e 1 - 1/ 2" NPT Nipple Conducti vity Cell P urg e Po rt C ompr es s ion H.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 4 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 Cable C onductivity Cell Ad ap t er * * Au t oma t ic Tempera t ur e Com pensato r , w hen specifie d by cat a log nu mber, is m ounted insid e ad apter.
Introduction 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 5 A D C B Ter mina l Boar d As s e m b l y #6- 32 Bi ndi ng He ad Sc r e w & Wa s h e r Vi e w of B a .
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 6 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99.
Specifications and Model Selection Guide 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7 2. Specifications and Model Selection Guide 2.1 Specifications Cell Constants 0.01, 0.1,1.0, 5, 10, 20, 25 and 50 cm -1 . Electrode Material Nickel or Platinum , as specified.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 8 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 Integral Automatic Temperature Compensation Refer to Section 2.2 – Model Selection Guid e Leadwire Tefzel covered, 18-g age cable (7 feet, 20 f eet, or Universal Head, as specified).
Specifications and Model Selection Guide 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 9 049 TABLE IV Selection 09 08 Leadwire Length ’ 7 ft.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 10 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 049 TABLE VII - OPTIONS Tagging None 0_ Linen L_ Stainless Steel S_ Certificate of Calibration No _ 0 Yes _ 1 Notes: 1.
Installation 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 11 3. Installation 3.1 Requirements To insur e that a re presenta tive sample is being measured at all times, the solutio n must m o ve thr ough and completely purge out the cell channels or guard tube.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 12 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 Allow at least 1/2 inch clearance bey ond the end of th e cell and 1/8 to 3/16 inch radius clearance surrounding the cell to p e rmit circulation of the solutio n.
Installation 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 13 3. Using a strap wrench on the m ounting nipple, tighten the assembly an additional 1-1/2 to 2 turns. Do not use stillson or chain type wrenches as they m ay damage and weaken the CPVC plastic.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 14 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 3.5 Removal 1. REDUCE PROCESS PRESSURE TO 50 PSIG OR LESS. Sh ut off purge line, if used. Dis connect wi ring conn ections if X1 is selected for Table IV in the MSG.
Installation 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 15 2.5" (64mm) App r ox 4.1" (105mm) App r ox 12.5" (317mm) X " See Bel ow App r ox 19.7" max . (502mm) Allow a minim um clearance o f 36" (914 m m) beyond this poin t for removal of electrode.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 16 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 2. 5" (64mm) Ap prox 4 .1" (105mm) Approx 12. 5" (317mm) X " See Below A pprox 19.
Maintenance and Replacement Parts 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 17 4. Maintenance and Replacement Parts If a series of below normal condu ctivity readings or abov e normal resistiv ity readings occu r, this m ay indicate that the cell is not filled with process solution.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 18 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 4.3 Replacement Parts Table 4-1 Replacement Parts Description Part Number Complete Ass embly See Figs.
Platinization and Platinum Black 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 19 5. Platinization and Platinum Black Only the electrodes havi ng cons tants from 5 t.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 20 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 A D C B View o f J unct ion Box Head w ith Cap Re moved WH I T E BLA CK RED GREEN Cell Temp. Com p.
Platinization and Platinum Black 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 21 A A D C B View o f J unct i on Box He ad with Cap Removed WH I T E BLA CK RED GREEN Cell Temp.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 22 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 WH I T E BLACK RED GREE N Cell Temp. Comp. WK S H G R Five Point Termin al Board. Each Terminal W i ll Accept #16 G age Max .
Platinization and Platinum Black 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 23 WH I T E BLA CK RED GREEN Cell Temp.
.
.
Industrial Auto mation and Co ntrol Honey well, Inc. 1100 Virginia D rive Fort Wa shington, Pennsy lvania 19034.
An important point after buying a device Honeywell 4909 CPVC (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Honeywell 4909 CPVC yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Honeywell 4909 CPVC - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Honeywell 4909 CPVC you will learn all the available features of the product, as well as information on its operation. The information that you get Honeywell 4909 CPVC will certainly help you make a decision on the purchase.
If you already are a holder of Honeywell 4909 CPVC, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Honeywell 4909 CPVC.
However, one of the most important roles played by the user manual is to help in solving problems with Honeywell 4909 CPVC. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Honeywell 4909 CPVC along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center