USER MANUAL H ELI O 2 HELIUM-OxygEn BLEndEr (DISS and NIST Connections) Model No. PM5400 Series (shown) PM5500 Series SA VE THESE INSTRUCTIONS CAUTION Federal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive T el: (+001) 610-262-6090 Northampton, P A 18067 USA Fax: (+001) 610-262-6080 www .
1 H ELI O 2 BLENDEr HELIUM-OXYGEN RECEIVING / INSPECTION Remove the Precision Medical, Inc. H ELI O 2 (Helium-Oxygen) Blender from the packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider . INTENDED USE The Precision Medical, Inc.
2 H ELI O 2 BLENDEr HELIUM-OXYGEN READ ALL INSTRUCTIONS BEFORE USING This manual instructs a Professional to install and operate the H ELI O 2 Blender . This is provided for your safety and to prevent damage to the H ELI O 2 Blender . If you do not understand this manual, DO NOT USE the H ELI O 2 Blender and contact your Provider .
3 H ELI O 2 BLENDEr HELIUM-OXYGEN W ARNING • Only trained, qualied medical personnel under the direct supervision of a licensed physician should operate the H ELI O 2 Blender . • Use this H ELI O 2 Blender only for its Intended Use as described in this manual.
4 H ELI O 2 BLENDEr HELIUM-OXYGEN SPECIFICA TIONS High Flow Low Flow Model PM5580: 80/20 heliox PM5570: 70/30 heliox PM5480: 80/20 heliox PM5470: 70/30 heliox Primary Outlet Flow Range 15 - 120 lpm 3 - 30 lpm W i th b ot h s u p p l y p re s s u r e s a t 50 p s i ( 3.
5 H ELI O 2 BLENDEr HELIUM-OXYGEN SPECIFICA TIONS continued Dimensions: (without ttings) Depth: 4.9 in (12.5 cm) Width: 2.3 in (5.7 cm) Height: 4.1 in (10.
6 H ELI O 2 BLENDEr HELIUM-OXYGEN DIAGRAMS CAUTION Missing or illegible labels must be replaced, contact Precision Medical, Inc. PM5400 and PM5500 Models A B F E D C H G.
7 H ELI O 2 BLENDEr HELIUM-OXYGEN COMPONENT DESCRIPTION ITEM DESCRIPTION A Oxygen Concentration Dial A dia l used for sele cti ng oxyg en con cen tra tion s betw een 20% -100% or 30% -100%. The F I O 2 scale is used for reference only . The actual F I O 2 must be veried with an Alarmed Oxygen Monitor / Analyzer .
8 H ELI O 2 BLENDEr HELIUM-OXYGEN PRE-USE TESTING W ARNING • Read this User Manual before installing or operating the H ELI O 2 Blender . • Conrm the concentration of heliox with an alarmed Oxygen Monitor / Analyzer . • Conrm contents of heliox cylinder prior to use.
9 H ELI O 2 BLENDEr HELIUM-OXYGEN REVERSE GAS FLOW PROCEDURE ( CE Requirements ONL Y) 1. Assure bleed ow valve is not engaged. Disconnect the oxygen hose from the gas source. Remove all outlet connections from the H ELI O 2 Blender to ensure that there is no outlet ow .
10 H ELI O 2 BLENDEr HELIUM-OXYGEN 7. Conrm the concentration of heliox with an Oxygen Monitor / Analyzer . If necessary activate the bleed ow valve to maintain F I O 2 accuracy . 8. T o activate the bleed, turn and rotate the knurled collar back until it contacts the cover .
1 1 H ELI O 2 BLENDEr HELIUM-OXYGEN OXYGEN FLOWMETER CONVERSIONS Oxygen Flowmeter Setting Corrected Heliox Flow (LPM) at V arious F I O 2 Settings 20% 30% 40% 50% 60% 70% 80% 90% 100% 1 1.8 1.6 1.4 1.3 1.2 1.18 1.15 1.02 1.0 2 3.6 3.2 2.8 2.6 2.4 2.4 2.
12 H ELI O 2 BLENDEr HELIUM-OXYGEN TECHNICAL DESCRIPTION For a complete T echnical Description of the H ELI O 2 Blender and list of Replacement Parts, reference the H ELI O 2 Blender Service Manual (P/N 506124) available on the Internet; www .precisionmedical.
13 H ELI O 2 BLENDEr HELIUM-OXYGEN TROUBLESHOOTING If the H ELI O 2 Blender fails to function, consult the T roubleshooting Guide below . If problem cannot be solved by using T roubleshooting Guide, refer to the H ELI O 2 Blender Service Manual (P/N 506124) available on the Internet; www .
14 H ELI O 2 BLENDEr HELIUM-OXYGEN LIMITED W ARRANTY AND LIMIT A TION OF LIABILITY Precis ion Medical, Inc. warran ts that the HeliO 2 Blende r , (the Product), will be free of defects in workmanship .
DECLARA TION OF CONFORMITY Manufacturer: Precision Medical, Inc. 300 Held Drive, Northampton, P A 18067, USA CONT ACT : Quality Manager Phone: 610-262-6090 Authorized European Representative: Emergo Europe (European Ofce) Molenstraat 15 2513 BH, The Hague The Netherlands Phone: +31 (0) 70.
An important point after buying a device Helio PM5400 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Helio PM5400 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Helio PM5400 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Helio PM5400 you will learn all the available features of the product, as well as information on its operation. The information that you get Helio PM5400 will certainly help you make a decision on the purchase.
If you already are a holder of Helio PM5400, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Helio PM5400.
However, one of the most important roles played by the user manual is to help in solving problems with Helio PM5400. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Helio PM5400 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center