6060 Spine Road Boulder, CO 80301 USA T 303 444 2009 F 303 444 9543 GE Analytical Instruments, Inc. © 2005H Page 1 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI Sievers 800 TOC Analyzer Flow Switch Upgrade (Flow Switch Kit and Firmware 2.12 CBI) 1.
GE Analytical Instruments, Inc. © 2005H Page 2 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI FIGURE ONE: Flow Switch Kit 4. Mechanical installation Refer to Figures One or Two for this proced ure. Reference will depend on orientation of tubing from regulator.
GE Analytical Instruments, Inc. © 2005H Page 3 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI 60 MICRON FILTER FLOW SWITCH REGULATOR SAMPLING BLOCK SWITCH OUTLET SWITCH BLOCK INLET INLET OUTLET FLOW SAMPLE REGULATOR FLOW REGULATOR INLET MODULE BINARY INPUT P2 FIGURE TWO: Installed Flow Switch on the Model 800 TOC Analyzer 5.
GE Analytical Instruments, Inc. © 2005H Page 4 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI TC ZERO CORRECTI ON -1.3250000 ENTER CLEAR ENT = SAVE CHAR/NUM CLR = RESTAR T/EXIT A RROWS = CHANGE CHA R FIGURE THREE: TOC ZERO OFFSET Screen 5.2.3.
GE Analytical Instruments, Inc. © 2005H Page 5 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI 5.2.9. Select each consumable, pre ss Enter. Write the information on the Printout or the Constants Re cord Form. Record date and percent left for the OXIDIZER.
GE Analytical Instruments, Inc. © 2005H Page 6 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI 5.3.4. Replace the EPROM from a Zero Insertion Force Socket. 5.3.4.1. There is a small brass lever on the socket to secure the EPROM chip. It should be in the locked position (end of the lever close to the PC board).
GE Analytical Instruments, Inc. © 2005H Page 7 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI carrier, gently press the pins of the chip to conform to the carrier. CAUTION: Be sure the NOTCH of the EPROM chip is toward the bottom of the analyzer and all the pins are engaged into th e socket.
GE Analytical Instruments, Inc. © 2005H Page 8 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI CAUTION: IF THERE ARE ANY VARIATIONS, CONSTANTS MUST BE ENTERED THROUGH THE FACTORY ACCESS MENU. ACCESS TO THIS MENU IS LIMITED AND WILL REQUIRE A PASSWORD.
GE Analytical Instruments, Inc. © 2005H Page 9 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI FIGURE NINE : First Constant Screen 5.6.1.8. Use the Constants Record Form as a guide. If a number is on the Constants Record Form, the new constant value must be that value.
GE Analytical Instruments, Inc. © 2005H Page 10 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI to save digits. After resetting the cl ock, Warning 35 will occur until the power has been cycled two times. 5.7.2. Set Flowrates – The Firmware allows selection of different flow rates for on- line sampling and grab (container) samp ling.
GE Analytical Instruments, Inc. © 2005H Page 11 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI Constants Record CONSTANT VALUE NAME CONSTANT VALUE AFTER UPGRADE CONSTANT VALUE PRIOR TO UPGRADE C1 CONST: Same as prior value C2 CONST: Same as prior value C1 THERM A: Same as prior value C2 THERM.
GE Analytical Instruments, Inc. © 2005H Page 12 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI CONSTANT VALUE NAME CONSTANT VALUE AFTER UPGRADE CONSTANT VALUE PRIOR TO UPGRADE ANALOG MAX: Sam.
GE Analytical Instruments, Inc. © 2005H Page 13 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI CONSTANT VALUE NAME CONSTANT VALUE AFTER UPGRADE CONSTANT VALUE PRIOR TO UPGRADE C1 XFER (50Hz): Same as prior value C2 XFER (50Hz): Same as prior value TC/IC OFFSET (50Hz): Same as prior value IC COND.
GE Analytical Instruments, Inc. © 2005H Page 14 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI 6. Firmware Configuration It is necessary to configure the binary inputs following the firmware upgrade. Follow these instructions to configure inputs and select proper polarity.
GE Analytical Instruments, Inc. © 2005H Page 15 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI ENABLE/DISABLE SET POLARITY ENTER CLEA R FIGURE THIRTEEN: Analyzer’s INPUT Menu 6.
GE Analytical Instruments, Inc. © 2005H Page 16 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI normally open or closed. Press Clear twice to return to the INPUT Menu. 6.2.8. Press Clear to return to the MAIN MENU. The analyzer will function as described in the Operation and Service Manual.
An important point after buying a device GE 800 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought GE 800 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data GE 800 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, GE 800 you will learn all the available features of the product, as well as information on its operation. The information that you get GE 800 will certainly help you make a decision on the purchase.
If you already are a holder of GE 800, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime GE 800.
However, one of the most important roles played by the user manual is to help in solving problems with GE 800. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device GE 800 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center