® PN 1554518 January 2000 ©2000 Fluke C orp oratio n. All rights reserv ed. Pri nted i n U.S .A. 1 Model 65 Infrared Thermometer Calibration Info rmation Introduct ion This calibr ation info rmation.
Model 65 Calibration Inf ormation 2 Safety In form ation A Warning To avoid possible personal injury: • The information provided in this manual is for the use of qualified personne l only. Do not perform the verification tests or calibration procedures described in this manual unless you a re qualified to do so.
Spec ific ati ons 3 Specifi catio ns Measuremen t Range − 40 °C to + 500 °C ( − 40 °F t o + 932 °F) Operating Temperatur e (Ambient) 0 °C to + 50 °C ( + 32 °F to + 122 °F) Storage Temperatur e − 20 °C to + 70 °C ( − 4 °F to + 158 °F), withou t battery Display Resoluti on Below 200 ° : 0.
Model 65 Calibration Inf ormation 4 Mainte nance Installing the Batt eries Remove th e holster to acces s the batt ery door. Figure 1 s hows how to rep lace the ba tteries. Caution • Match the + and − polarities of the battery wit h the batt ery case.
Required Eq uip men t 5 Require d Equipme nt Table 1 shows the equ ipmen t required for the performan ce tes ts and cal ibration a djust ments. Table 1. Requir ed Equipment Equipment Minimum Specific ations Recomm ended Model Blackbody Calibratio n Source Temperatur e range: Am bient to 4 00 °C Minimum apertur e diam eter: 76 mm (3.
Model 65 Calibration Inf ormation 6 Table 2. Performance Test Values Temperature at 0.95 Emissivity UUT Reading Limits 1, 2 Room temperatur e Room temperatur e ± 2 °C ( ± 3.6 °F) 200 °C (392 °F) 196 °C to 204 °C (384.8 °F to 3 99.2 °F) 300 °C (572 °F) 294 °C to 306 °C (561.
Calibrat ion Ad ju stm ent s 7 MIN MAX ° C ° F Room temperature calibration mode High temperature calibration mode End calibration mode Press to exit calibration mode F actory Calibration mode. Caution Do not change the factory calibration setting acd01f.
Model 65 Calibration Inf ormation 8 8. Press D to enter the high -temperatu re adju stmen t mode (disp lay shows 2_H ), as shown in Fi gure 2. 9. Press G .
An important point after buying a device Fluke 65 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Fluke 65 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Fluke 65 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Fluke 65 you will learn all the available features of the product, as well as information on its operation. The information that you get Fluke 65 will certainly help you make a decision on the purchase.
If you already are a holder of Fluke 65, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Fluke 65.
However, one of the most important roles played by the user manual is to help in solving problems with Fluke 65. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Fluke 65 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center