T emporal Scanner TM 2000C EX ER GE N Quick Check-List: - Read instructions completely before using - See www .temporalscanner .com for Educational Video - Remove protective cap before using - Reads arterial temperature, which is a core temperature very close to rect al temperature (See pg.
READ ALL INSTRUCTIONS BEFORE USING When using the product, especially when children are present, basic safety precautions should always be followed, including the following: z This product is intended for household use only . For information on thermometers for professional use, please see www .
z If your thermometer will not be used regularly , remove the battery to prevent possible damage due to chemical leakage. If the battery leaks, remove carefully . Do not allow bare skin to touch leaking fluid. z Dispose of used batteries properly . Do not wrap them in metal or aluminum foil.
The T emporalScanner has patented software, providing arterial heat balance. This unique process determines temperature by accurately measuring the balance between the tissues warming from arterial blood and tissues cooling/ warming caused by heat loss/gain to the environment.
5 Probe Cone Probe Sensor Lens Probe LCD Display Screen LED Indicator Light SCAN Button Automatic turn-off in 30 seconds LED Indicator Light Thumb-Push Ridges Battery Comp artment 9-volt alkaline battery z T o Scan: Depress the button. The instrument will continually scan for the highest temperature (peak) as long as the button is depressed.
What you should know before using the T A T : z Measure only the side of the head exposed to the environment. Anything covering the area to be measured (hair , hat, wig, bandages) would insulate the area, resulting in falsely high readings. z Slide the thermometer straight across the forehead, not down the side of the face.
T o ensure that the reading always reflects the body temperature accurately , you need to take account of the following factors which may affect an accurate reading. Sweating: When a fever resolves, your body may bring its temperature down by sweating.
How to T ake a T emperature 1. Remove protective cap before use. Be sure lens is clean. If not, clean with cotton swab dipped in alcohol and let dry . Hold the ther- mometer as shown . 2. Gently position the probe flush (flat) on the center of the forehead, midway between the eyebrow and the hairline.
Normal Body T emperature (BT) Normal BT is not a single temperature, but a range of temperatures influ- enced by age, time of day , and measurement site. Y ou can establish your family’ s normal ranges by taking a number of temperatures from each mem- ber during a day and keeping records of them.
On a stable resting individual, temporal artery temperature is about the same as a rectal temperature, and approximately 0.8°F (0.4°C) higher than an opti- mum oral temperature.
Rectal T emperature Artifacts: Rectal temperature should only be con- sidered as a good approximation of core temperature when the patient’s thermal balance is stable, 8 and may be misleading after antipyretics, physical exercise, or other intervention that may change temperature quickly .
Cleaning the Instrument The T emporalScanner is an optical instrument. Like a camera or eye glasses, a dirty lens will distort the view . If the thermometer is unable to see the heat clearly , it will be unable to measure it accurately , resulting in low readings.
Changing the Battery Blinking battery icon with temperature displayed: battery is low but will still operate correctly . Replace soon. Blinking battery icon with 2 dashes: not enough energy in the battery to measure correct temperature. Replace battery .
Display Messages A flickering Scn on display is visible during measurement. At completion, releasing the button will display and lock temperature on the screen for 30 seconds. The target temperature measured is higher than 107.6°F (42°C). The target temperature measured is lower than 60°F (15.
Clinical Accuracy Meets ASTM E1965-98 and EN60601-1 standards for electronic and radiation thermometers to the extent applicable to thermometers which measure the surface of the skin over the temporal artery . Regulatory Approvals CE Mark to -0197, TUV , Declaration of Conformity-ISO 9003/08.
One Y ear W arranty Exergen Corporation warrants each new Exergen T emporalScanner 2000C (except battery) against defects in materials or workmanship for a period of one year from the date of purchase, and agrees to repair or replace any defective product without charge.
An important point after buying a device Exergen Temporal Scanner 2000C (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Exergen Temporal Scanner 2000C yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Exergen Temporal Scanner 2000C - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Exergen Temporal Scanner 2000C you will learn all the available features of the product, as well as information on its operation. The information that you get Exergen Temporal Scanner 2000C will certainly help you make a decision on the purchase.
If you already are a holder of Exergen Temporal Scanner 2000C, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Exergen Temporal Scanner 2000C.
However, one of the most important roles played by the user manual is to help in solving problems with Exergen Temporal Scanner 2000C. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Exergen Temporal Scanner 2000C along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center