Instruction/ maintenance manual of the product MMI-20019043 Emerson
Go to page of 374
Configuration and Use Manual MMI-20019043, Rev AA March 2012 Micro Motion ® Model 2700 Transmitters with Analog Outputs Configuration and Use Manual.
Safety messages Safety messages are provided throughout this manual to protect personnel and equipment. Read each safety message carefully before proceeding to the next step. Micro Motion customer service Location Telephone number Email U.S.A. 800-522-MASS (800-522-6277) (toll free) flow.
Contents Part I Getting Started Chapter 1 Before you begin .............................................................................................................3 1.1 About this manual ...........................................................
4.5 Configure density measurement ................................................................................................. 45 4.5.1 Configure Density Measurement Unit ...........................................................................
5.6.6 Configure Sensor Liner Material .................................................................................... 97 5.6.7 Configure Sensor Flange Type ....................................................................................... 98 Chapter 6 Integrate the meter with the control system .
9.4.3 View and acknowledge alerts using ProLink III ............................................................. 156 9.4.4 View alarms using the Field Communicator ................................................................ 156 9.4.5 Alarm data in transmitter memory .
12.3 Flow measurement problems .................................................................................................... 220 12.4 Density measurement problems .................................................................................
B.2.2 Make a service port connection ................................................................................... 275 B.2.3 Make a HART/Bell 202 connection .............................................................................. 276 B.2.4 Make a HART/RS-485 connection .
Part I Getting Started Chapters covered in this part: • Before you begin • Quick start Getting Started Configuration and Use Manual 1.
Getting Started 2 Micro Motion ® Model 2700 Transmitters with Analog Outputs.
1 Before you begin Topics covered in this chapter: • About this manual • Transmitter model code • Communications tools and protocols • Additional documentation and resources 1.1 About this manual This manual provides information to help you configure, commission, use, maintain, and troubleshoot the Micro Motion Model 2700 transmitter.
Communications tools, protocols, and related information Table 1-1: Communica- tions tool Supported protocols Scope In this manual For more information Display Not applicable Basic configuration and commissioning Complete user informa- tion. See Appendix A .
Additional documentation and resources (continued) Table 1-2: Topic Document Transmitter installation Micro Motion Model 1700 and Model 2700 Transmitters: Installation Manual Hazardous area installa- tion See the approval documentation shipped with the transmitter, or download the appropriate documentation from the Micro Motion web site at www.
Before you begin 6 Micro Motion ® Model 2700 Transmitters with Analog Outputs.
2 Quick start Topics covered in this chapter: • Power up the transmitter • Check flowmeter status • Make a startup connection to the transmitter • Characterize the flowmeter (if required) • Verify mass flow measurement • Verify the zero 2.
1. Wait approximately 10 seconds for the power-up sequence to complete. Immediately after power-up, the transmitter runs through diagnostic routines and checks for error conditions. During the power-up sequence, Alarm A009 is active. This alarm should clear automatically when the power-up sequence is complete.
2.3 Make a startup connection to the transmitter For all configuration tools except the display, you must have an active connection to the transmitter to configure the transmitter. Follow this procedure to make your first connection to the transmitter.
2.4 Characterize the flowmeter (if required) Display Not available ProLink II • ProLink > Configuration > Device > Sensor Type • ProLink > Configuration > Flow • ProLink > Conf.
• Older curved-tube sensors (all sensors except T-Series): see Figure 2-1 • Newer curved-tube sensors (all sensors except T-Series): see Figure 2-2 • Older straight-tube sensors (T-Series): see .
Tag on older straight-tube sensor (T-Series) Figure 2-3: Tag on newer straight-tube sensor (T-Series) Figure 2-4: Density calibration parameters (D1, D2, K1, K2, FD, DT, TC) If your sensor tag does not show a D1 or D2 value: • For D1 , enter the Dens A or D1 value from the calibration certificate.
If your sensor tag does not show a DT or TC value, enter the last 3 digits of the density calibration factor. In the sample tag, this value is shown as 4.44 (see Figure 2-1 ). Flow calibration parameters (FCF, FT) Two separate values are used to describe flow calibration: a 6-character FCF value and a 4- character FT value.
• Review the troubleshooting suggestions for flow measurement issues. See Section 12.3 . 2.6 Verify the zero Verifying the zero helps you determine if the stored zero value is appropriate to your installation, or if a field zero can improve measurement accuracy.
3. If the zero verification procedure fails: a. Confirm that the sensor is completely blocked in, that flow has stopped, and that the sensor is completely full of process fluid. b. Verify that the process fluid is not flashing or condensing, and that it does not contain particles that can settle out.
3. If the zero verification procedure fails: a. Confirm that the sensor is completely blocked in, that flow has stopped, and that the sensor is completely full of process fluid. b. Verify that the process fluid is not flashing or condensing, and that it does not contain particles that can settle out.
Part II Configuration and commissioning Chapters covered in this part: • Introduction to configuration and commissioning • Configure process measurement • Configure device options and preference.
Configuration and commissioning 18 Micro Motion ® Model 2700 Transmitters with Analog Outputs.
3 Introduction to configuration and commissioning Topics covered in this chapter: • Configuration flowchart • Default values and ranges • Enable access to the off-line menu of the display • Disable write-protection on the transmitter configuration • Restore the factory configuration 3.
Configuration flowchart Figure 3-1: Integrate device with control system Configure device options and preferences Configure process measurement Configure mass flow measurement Configure volume flow me.
3.2 Default values and ranges See Section E.1 to view the default values and ranges for the most commonly used parameters. 3.3 Enable access to the off-line menu of the display Display OFF-LINE MAINT .
Tip Write-protecting the transmitter prevents accidental changes to configuration. It does not prevent normal operational use. You can always disable write-protection, perform any required configuration changes, then re-enable write-protection.
4 Configure process measurement Topics covered in this chapter: • Configure mass flow measurement • Configure volume flow measurement for liquid applications • Configure gas standard volume (GSV.
Tip If the measurement unit you want to use is not available, you can define a special measurement unit. Options for Mass Flow Measurement Unit The transmitter provides a standard set of measurement units for Mass Flow Measurement Unit , plus one user-defined special measurement unit.
Define a special measurement unit for mass flow Display Not available ProLink II ProLink > Configuration > Special Units ProLink III Device Tools > Configuration > Process Measurement >.
a. 1 lb/sec = 16 oz/sec b. Mass Flow Conversion Factor = 1/16 = 0.0625 4. Set Mass Flow Conversion Factor to 0.0625 . 5. Set Mass Flow Label to oz/sec .
• In general, lower damping values are preferable because there is less chance of data loss, and less lag time between the actual measurement and the reported value. • For gas applications, Micro Motion recommends setting Flow Damping to 2.56 or higher.
Procedure Set Mass Flow Cutoff to the value you want to use. The default value for Mass Flow Cutoff is 0.0 g/sec or a sensor-specific value set at the factory. The recommended setting is 0.05% of the sensor's rated maximum flow rate or a value below the highest expected flow rate.
- The frequency output will report the actual flow rate, and the actual flow rate will be used in all internal processing. • If the mass flow rate drops below 10 g/sec, both outputs will report zero flow, and 0 will be used in all internal processing.
Procedure Set Volume Flow Type to Liquid . 4.2.2 Configure Volume Flow Measurement Unit for liquid applications Display OFF-LINE MAINT > OFF-LINE CONFG > UNITS > VOL ProLink II ProLink > C.
Options for Volume Flow Measurement Unit for liquid applications (continued) Table 4-3: Unit description Label Display ProLink II ProLink III Field Communica- tor Cubic feet per hour CUFT/H ft3/hr ft3.
Define a special measurement unit for volume flow Display Not available ProLink II ProLink > Configuration > Special Units ProLink III Device Tools > Configuration > Process Measurement &g.
a. 1 gal/sec = 8 pints/sec b. Volume Flow Conversion Factor = 1/8 = 0.1250 4. Set Volume Flow Conversion Factor to 0.1250 . 5. Set Volume Flow Label to pints/sec .
Result: If the volume flow rate drops below 15 l/sec, volume flow will be reported as 0, and 0 will be used in all internal processing. Example: Cutoff interaction with AO Cutoff higher than Volume Fl.
4.3.1 Configure Volume Flow Type for gas applications Display Not available ProLink II ProLink > Configuration > Flow > Vol Flow Type ProLink III Device Tools > Configuration > Process .
Procedure Set Standard Gas Density to the standard reference density of the gas you are measuring. Note ProLink II and ProLink III provide a guided method that you can use to calculate the standard density of your gas, if you do not know it.
Options for Gas Standard Volume Measurement Unit Table 4-4: Unit description Label Display ProLink II ProLink III Field Communica- tor Normal cubic meters per sec- ond NM3/S Nm3/sec Nm3/sec Nm3/sec No.
Overview A special measurement unit is a user-defined unit of measure that allows you to report process data, totalizer data, and inventory data in a unit that is not available in the transmitter. A special measurement unit is calculated from an existing measurement unit using a conversion factor.
4.3.4 Configure Gas Standard Volume Flow Cutoff Display Not available ProLink II ProLink > Configuration > Flow > Std Gas Vol Flow Cutoff ProLink III Device Tools > Configuration > Proc.
Example: Cutoff interaction with AO Cutoff higher than Gas Standard Volume Flow Cutoff Configuration: • mA Output Process Variable for the primary mA output: Gas Standard Volume Flow Rate • Freque.
4.4.1 Options for Flow Direction Options for Flow Direction Table 4-5: Flow Direction setting Relationship to Flow Direction ar- row on sensor ProLink II ProLink III Field Communicator Forward Forward Forward Appropriate when the Flow Direction arrow is in the same direction as the majority of flow.
Effect of Flow Direction on the mA output: Lower Range Value = 0 Figure 4-1: Flow Direction = Forward mA output -x 0 x Reverse flow Forward flow 20 12 4 Flow Direction = Reverse, Negate Forward mA out.
• Under conditions of reverse flow or zero flow, the mA output is 4 mA. • Under conditions of forward flow, up to a flow rate of 100 g/sec, the mA output varies between 4 mA and 20 mA in proportion to the flow rate.
Effect of Flow Direction on frequency outputs Flow Direction affects how the transmitter reports flow values via the frequency outputs. The frequency outputs are affected by Flow Direction only if Frequency Output Process Variable is set to a flow variable.
Effect of the Flow Direction parameter and actual flow direction on flow values reported via digital communications Table 4-8: Flow Direction setting Actual flow direction Forward Zero flow Reverse Fo.
4.5.1 Configure Density Measurement Unit Display OFF-LINE MAINT > OFF-LINE CONFG > UNITS > DENS ProLink II ProLink > Configuration > Density > Density Units ProLink III Device Tools .
4.5.2 Configure slug flow parameters Display Not available ProLink II • ProLink > Configuration > Density > Slug High Limit • ProLink > Configuration > Density > Slug Low Limit .
The default value for Slug High Limit is 5.0 g/cm 3 . The range is 0.0 to 10.0 g/cm 3 . 3. Set Slug Duration to the number of seconds that the transmitter will wait for a slug flow condition to clear before performing the configured slug flow action. The default value for Slug Duration is 0.
Overview Damping is used to smooth out small, rapid fluctuations in process measurement. Damping Value specifies the time period (in seconds) over which the transmitter will spread changes in the reported process variable. At the end of the interval, the reported process variable will reflect 63% of the change in the actual measured value.
Interaction between Density Damping and Added Damping In some circumstances, both Density Damping and Added Damping are applied to the reported density value. Density Damping controls the rate of change in the density process variable. Added Damping controls the rate of change reported via the mA output.
4.6.1 Configure Temperature Measurement Unit Display OFF-LINE MAINT > OFF-LINE CONFG > UNITS > TEMP ProLink II ProLink > Configuration > Temperature > Temp Units ProLink III Device T.
Overview Damping is used to smooth out small, rapid fluctuations in process measurement. Damping Value specifies the time period (in seconds) over which the transmitter will spread changes in the reported process variable. At the end of the interval, the reported process variable will reflect 63% of the change in the actual measured value.
4.7 Configure the petroleum measurement application The petroleum measurement application enables Correction for the effect of Temperature on the volume of Liquids (CTL), by calculating and applying a Volume Correction Factor (VCF) to volume measurement.
Option Setup Polling for tempera- ture a. Ensure that the primary mA output has been wired to support HART polling. b. Choose View > Preferences . c. Enable Use External Temperature . d. Choose ProLink > Configuration > Polled Variables . e. Choose an unused polling slot.
5. Set Temperature Source to the method that the transmitter will use to obtain temperature data. Option Description Poll for external value The transmitter will poll an external temperature device, us- ing HART protocol over the primary mA output. RTD The transmitter will use the temperature data from the sen- sor.
4.7.3 Configure petroleum measurement using the Field Communicator 1. Choose Online > Configure > Manual Setup > Measurements > Set Up Petroleum . 2. Specify the API table to use. a. Open the Petroleum Measurement Source menu and select the API table number.
Option Setup A value written by digital communica- tions a. Choose Online > Configure > Manual Setup > Measurements > External Pressure/Temperature > Temperature .
API reference tables, associated process fluids, and associated calculation values (continued) Table 4-13: Table name Process fluid CTL source data Reference temperature Density unit 53B Generalized p.
• The concentration measurement application must be enabled on your transmitter. • The concentration matrix you want to use must be available on your transmitter, or it must be available as a file on your computer. • You must know the derived variable that your matrix is designed for.
Restriction The high and low limit alarms require the enhanced core processor. Example: If Alarm Limit is set to 5%, Enable Temp High is checked, and the matrix is built for a temperature range of 40 °F to 80 °F, an extrapolation alarm will be posted if process temperature goes above 82 °F 7.
Option Setup A value written by digital communica- tions a. Choose View > Preferences . b. Enable Use External Temperature . c. Perform the necessary host programming and communications setup to write temperature data to the transmitter at appropri- ate intervals.
3. Choose Device Tools > Configuration > Process Measurement > Concentration Measurement . 4. Set Derived Variable to the derived variable that your matrix is designed for, and click Apply . Important • All concentration matrices on your transmitter must use the same derived variable.
Restriction The high and low limit alarms require the enhanced core processor. Example: If Extrapolation Alarm Limit is set to 5%, High Extrapolation Limit (Temperature) is enabled, and the matrix is built for a temperature range of 40 °F to 80 °F, an extrapolation alarm will be posted if process temperature goes above 82 °F 8.
13. Set Active Matrix to the matrix to be used for measurement. Concentration process variables are now available on the transmitter. You can view and report them in the same way that you view and report other process variables.
d. Choose Online > Configure > Alert Setup > CM Alerts . e. Enable or disable the high and low limit alarms for temperature and density, as desired.
Option Setup Polling for tempera- ture a. Ensure that the primary mA output has been wired to support HART polling. b. Choose Online > Configure > Manual Setup > Measurements > External Pressure/Temperature > Temperature . c. Enable External Temperature .
Tip If the standard matrices are not appropriate for your application, you can build a custom matrix or purchase a custom matrix from Micro Motion. Standard concentration matrices and associated measu.
Derived variables and calculated process variables Table 4-15: Derived Variable Description Calculated process variables Density at reference tempera- ture Standard volume flow rate Specific gravity C.
Derived variables and calculated process variables (continued) Table 4-15: Derived Variable Description Calculated process variables Density at reference tempera- ture Standard volume flow rate Specif.
The flow factor is the percent change in the flow rate per PSI. When entering the value, reverse the sign. Example: If the flow factor is 0.000004 % per PSI, enter − 0.000004 % per PSI. 4. Enter Density Factor for your sensor. The density factor is the change in fluid density, in g/cm 3 /PSI.
Option Setup A value written by digital communica- tions a. Set Pressure Units to the desired unit. b. Perform the necessary host programming and communications setup to write pressure data to the transmitter at appropriate in- tervals.
The density factor is the change in fluid density, in g/cm 3 /PSI. When entering the value, reverse the sign. Example: If the density factor is 0.000006 g/cm 3 /PSI, enter − 0.000006 g/cm3/PSI. 6. Set Pressure Source to the method that the transmitter will use to obtain pressure data.
9. If you want to use digital communications, click Apply , then perform the necessary host programming and communications setup to write temperature data to the transmitter at appropriate intervals.
Option Setup A user-configured static pressure val- ue a. Set Pressure Unit to the desired unit. b. Set Compensation Pressure to the desired value. Polling for pressure a.
Options for Pressure Measurement Unit Table 4-16: Unit description Label Display ProLink II ProLink III Field Communica- tor Feet water @ 68 °F FTH2O Ft Water @ 68°F Ft Water @ 68°F ftH2O Inches wa.
Configure process measurement 76 Micro Motion ® Model 2700 Transmitters with Analog Outputs.
5 Configure device options and preferences Topics covered in this chapter: • Configure the transmitter display • Enable or disable operator actions from the display • Configure security for the display menus • Configure response time parameters • Configure alarm handling • Configure informational parameters 5.
The languages available depend on your transmitter model and version. 5.1.2 Configure the process variables shown on the display Display Not available ProLink II ProLink > Configuration > Displa.
Display variable Process variable assignment Display Variable 10 None Display Variable 11 None Display Variable 12 None Display Variable 13 None Display Variable 14 None Display Variable 15 None Confi.
Overview Setting Display Precision determines the precision (number of decimal places) shown on the display. You can set Display Precision independently for each variable. Setting Display Precision does not affect the actual value of the process variable.
5.1.5 Enable or disable automatic scrolling through the display variables Display OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > AUTO SCRLL ProLink II ProLink > Configuration > Display >.
Procedure Enable or disable Backlight . The default setting is Enabled . 5.1.7 Enable or disable Status LED Blinking Display Not available ProLink II ProLink > Configuration > Display > Displ.
5.2.1 Enable or disable Totalizer Start/Stop from the display Display OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > TOTALS STOP ProLink II ProLink > Configuration > Display > Display O.
Restrictions • This parameter does not apply to inventories. You cannot reset inventories from the display. • You cannot use the display to reset all totalizers as a group.
Option Description Enabled (default) Operators can use a single display command to acknowledge all alarms at once. Disabled Operators cannot acknowledge all alarms at once, they must be ac- knowledged individually.
3. To require a password for access to the maintenance section of the off-line menu and the Smart Meter Verification menu, enable or disable Off-Line Password .
5.4.1 Configure Update Rate Display Not available ProLink II ProLink > Configuration > Device > Update Rate ProLink III Device Tools > Configuration > Process Measurement > Response .
Effects of Update Rate = Special Incompatible features and functions Special mode is not compatible with the following features and functions: • Enhanced events. Use basic events instead. • All calibration procedures. • Zero verification. • Restoring the factory zero or the prior zero.
5.4.2 Configure Calculation Speed (Response Time) Display Not available ProLink II ProLink > Configuration > Device > Response Time ProLink III Device Tools > Configuration > Process Me.
5.5.1 Configure Fault Timeout Display Not available ProLink II ProLink > Configuration > Analog Output > Last Measured Value Timeout ProLink > Configuration > Frequency/Discrete Output .
Overview Use Status Alarm Severity to control the fault actions that the transmitter performs when it detects an alarm condition. Restrictions • For some alarms, Status Alarm Severity is not configurable. • For some alarms, Status Alarm Severity can be set only to two of the three options.
Status alarms and Status Alarm Severity Table 5-2: Alarm code Status message Default severity Notes Configurable? A001 EEPROM Error (Core Pro- cessor) Fault No A002 RAM Error (Core Processor) Fault No.
Status alarms and Status Alarm Severity (continued) Table 5-2: Alarm code Status message Default severity Notes Configurable? A031 Low Power Fault Applies only to flowmeters with the enhanced core processor. No A032 Meter Verification in Pro- gress: Outputs to Fault Varies Applies only to transmitters with Smart Meter Verification.
Status alarms and Status Alarm Severity (continued) Table 5-2: Alarm code Status message Default severity Notes Configurable? A114 mA Output 2 Fixed Informational Can be set to either Informational or Ignore , but cannot be set to Fault .
- Message - Date • Sensor parameters - Sensor Serial Number - Sensor Material - Sensor Liner Material - Sensor Flange Type 5.6.1 Configure Descriptor Display Not available ProLink II ProLink > Co.
5.6.3 Configure Date Display Not available ProLink II ProLink > Configuration > Device > Date ProLink III Device Tools > Configuration > Informational Parameters > Transmitter Field .
5.6.5 Configure Sensor Material Display Not available ProLink II ProLink > Configuration > Sensor > Sensor Matl ProLink III Device Tools > Configuration > Informational Parameters > .
5.6.7 Configure Sensor Flange Type Display Not available ProLink II ProLink > Configuration > Sensor > Flange ProLink III Device Tools > Configuration > Informational Parameters > Se.
6 Integrate the meter with the control system Topics covered in this chapter: • Configure the transmitter channels • Configure the mA output • Configure the frequency output • Configure the discrete output • Configure events • Configure digital communications 6.
Postrequisites For each channel that you configured, perform or verify the corresponding input or output configuration. When the configuration of a channel is changed, the channel’s behavior will be.
• If you plan to configure an output to report a concentration measurement process variable, ensure that the concentration measurement application is configured so that the desired variable is available.
Options for mA Output Process Variable (continued) Table 6-1: Process variable Label Display ProLink II ProLink III Field Communicator Concentration measurement Density at reference RDENS CM: Density .
Procedure Set LRV and URV as desired. • LRV is the value of mA Output Process Variable represented by an output of 4 mA. The default value for LRV depends on the setting of mA Output Process Variable . Enter LRV in the measurement units that are configured for mA Output Process Variable .
Default values for Lower Range Value (LRV) and Upper Range Value (URV) (continued) Table 6-2: Process variable LRV URV Concentration 0% 100% Baume 0 10 Specific gravity 0 10 6.
Example: Cutoff interaction Configuration: • mA Output Process Variable = Mass Flow Rate • Frequency Output Process Variable = Mass Flow Rate • AO Cutoff = 10 g/sec • Mass Flow Cutoff = 15 g/sec Result: If the mass flow rate drops below 15 g/sec, all outputs representing mass flow will report zero flow.
Note Added Damping is not applied if the mA output is fixed (for example, during loop testing) or if the mA output is reporting a fault. Added Damping is applied while sensor simulation is active. Procedure Set Added Damping to the desired value. The default value is 0.
Result: A change in the mass flow rate will be reflected in the mA output over a time period that is greater than 3 seconds. The exact time period is calculated by the transmitter according to internal algorithms which are not configurable.
Options for mA Output Fault Action and mA Output Fault Level (continued) Table 6-4: Option mA output behavior mA Output Fault Level None Tracks data for the assigned process vari- able; no fault actio.
Overview Frequency Output Process Variable controls the variable that is reported over the frequency output. Prerequisites If you plan to configure the output to report volume flow, ensure that you have set Volume Flow Type as desired: Liquid or Gas Standard Volume .
Options for Frequency Output Process Variable (continued) Table 6-5: Process variable Label Display ProLink II ProLink III Field Communica- tor Net mass flow NET M ED: Net Mass Flow Rate Net Mass Flow Rate ED Net Mass flo Net volume flow NET V ED: Net Vol Flow Rate Net Volume Flow Rate ED Net Vol flo 6.
6.3.3 Configure Frequency Output Scaling Method Display OFF-LINE MAINT > OFF-LINE CONFG > IO > CH B > SET FO > FO SCALE ProLink II ProLink > Configuration > Frequency/Discrete Out.
N Number of pulses per flow unit, as configured in the receiving device The resulting Frequency Factor must be within the range of the frequency output (0 to 10,000 Hz): • If Frequency Factor is less than1 Hz,reconfigure the receiving device for a higher pulses/unit setting.
The ON signal may be the high voltage or 0.0 V , depending on Frequency Output Polarity . Interaction of Frequency Output Maximum Pulse Width and Frequency Output Polarity Table 6-7: Polarity Pulse width Active High Active Low Procedure Set Frequency Output Maximum Pulse Width as desired.
Note For some faults only: If Last Measured Value Timeout is set to a non-zero value, the transmitter will not implement the fault action until the timeout has elapsed. Procedure 1. Set Frequency Output Fault Action as desired. The default value is Downscale (0 Hz).
• Discrete Output Polarity • Discrete Output Fault Action Restriction Before you can configure the discrete output, you must configure a channel to operate as a discrete output. Important Whenever you change a discrete output parameter, verify all other discrete output parameters before returning the flowmeter to service.
Options for Discrete Output Source (continued) Table 6-9: Option Label Condition Discrete out- put voltage Display ProLink II ProLink III Field Commu- nicator Enhanced Event 4 Enhanced Event 5 OFF 0 V.
Configure Flow Switch parameters Display OFF-LINE MAINT > OFF-LINE CONFG > IO > CH B > SET DO > CONFIG FL SW ProLink II • ProLink > Configuration > Flow > Flow Switch Variabl.
6.4.2 Configure Discrete Output Polarity Display OFF-LINE MAINT > OFF-LINE CONFG > IO > CH B > SET DO > DO POLAR ProLink II ProLink > Configuration > Frequency/Discrete Output >.
Typical discrete output circuit Figure 6-1: A. 24 V (Nom) B. 3.2 K Ω C. Out+ D. Out − 6.4.3 Configure Discrete Output Fault Action Display Not available ProLink II ProLink > Configuration > .
Procedure Set Discrete Output Fault Action as desired. The default setting is None . Options for Discrete Output Fault Action Options for Discrete Output Fault Action Table 6-11: Label Discrete output.
6.5.1 Configure a basic event Display Not available ProLink II ProLink > Configuration > Events ProLink III Device Tools > Configuration > Events > Basic Events Field Communicator Not available Overview A basic event is used to provide notification of process changes.
Overview An enhanced event is used to provide notification of process changes and, optionally, to perform specific transmitter actions if the event occurs.
Options for Enhanced Event Action Options for Enhanced Event Action Table 6-12: Action Label Display ProLink II ProLink III Field Communicator Standard None (default) NONE None None None Start sensor .
6.6 Configure digital communications The digital communications parameters control how the transmitter will communicate using digital communications. Your transmitter supports the following types of d.
2. Ensure Loop Current Mode ( mA Output Action ) is configured appropriately. Options Description Enabled The primary mA output reports process data as configured.
Label Description ProLink II ProLink III Field Communi- cator Primary Variable Source (Primary Variable) PV The transmitter sends the primary variable (PV) in the configured measurement units in each burst (e.
Options for HART variables Options for HART variables Table 6-13: Process variable Primary Varia- ble (PV) Secondary Variable (SV) Third Variable (TV) Fourth Varia- ble (QV ) Standard Mass flow rate .
Options for HART variables (continued) Table 6-13: Process variable Primary Varia- ble (PV) Secondary Variable (SV) Third Variable (TV) Fourth Varia- ble (QV ) ED standard volume flow rate ✓ ✓ ✓.
6.6.2 Configure HART/RS-485 communications Display OFF-LINE MAINT > OFF-LINE CONFG > COMM ProLink II ProLink > Configuration > Device > Digital Comm Settings > HART Address ProLink &.
6.6.3 Configure Modbus/RS-485 communications Display OFF-LINE MAINT > OFF-LINE CONFG > COMM ProLink II ProLink > Configuration > Device > Digital Comm Settings ProLink III Device Tools .
4. Set Parity , Stop Bits , and Baud Rate as appropriate for your network. 5. Set Floating-Point Byte Order to match the byte order used by your Modbus host. Code Byte order 0 1–2 3–4 1 3–4 1–2 2 2–1 4–3 3 4–3 2–1 See Table 6-15 for the bit structure of bytes 1, 2, 3, and 4.
Overview Digital Communications Fault Action specifies the values that will be reported via digital communications if the transmitter encounters an internal fault condition. Procedure Set Digital Communications Fault Action as desired. The default setting is None .
CAUTION! If you set mA Output Fault Action or Frequency Output Fault Action to None , be sure to set Digital Communications Fault Action to None . If you do not, the output will not report actual process data, and this may result in measurement errors or unintended consequences for your process.
Integrate the meter with the control system 134 Micro Motion ® Model 2700 Transmitters with Analog Outputs.
7 Completing the configuration Topics covered in this chapter: • Test or tune the system using sensor simulation • Back up transmitter configuration • Enable write-protection on the transmitter configuration 7.
Option Required values Sine Period Minimum Maximum 4. For density, set Wave Form as desired and enter the required values. Option Required values Fixed Fixed Value Sawtooth Period Minimum Maximum Sine Period Minimum Maximum 5. For temperature, set Wave Form as desired and enter the required values.
• All mass flow rate, temperature, and density values shown on the display or reported via outputs or digital communications • The mass total and mass inventory values • All volume calculations .
7.3 Enable write-protection on the transmitter configuration Display OFF-LINE MAINT > CONFIG > LOCK ProLink II ProLink > Configuration > Device > Enable Write Protection ProLink III Dev.
8 Set up the Weights & Measures application Topics covered in this chapter: • Weights & Measures application • Set up the Weights & Measures application using ProLink II • Set up the.
Metrological security Metrological security protects the transmitter from all changes that would affect measurement. This includes changes to configuration and some maintenance procedures. Micro Motion implements metrological security via “software security.
3. Set Approval to the appropriate regulatory agency for your application. Option Description NTEP Regulatory agency for the U.S.A. and Canada OIML Regulatory agency for all other world areas 4.
FVZ is a diagnostic variable that monitors the zero value over a period of 3 minutes. It must be read during meter commissioning to comply with MID requirements for Weights & Measures applications in Measuring Instrument Directive (MID) 2004/22/ EC.
8.3 Set up the Weights & Measures application using ProLink III When your transmitter is set up to comply with OIML or NTEP requirements, and the transmitter is “secured,” selected measurement data from the transmitter is approved for Weights & Measures applications.
“Digital communications” refers to any method that uses Modbus or HART communications to interact with the transmitter. This includes ProLink II, ProLink III, the Field Communicator, and any host. 4. If required for your installation, configure your transmitters for two frequency outputs and set them to operate in Quadrature mode.
9. Install the physical seal. Important In most installations, the physical seal is a wire seal that must be installed by a certified Weights & Measures inspector. The seal is provided by the inspector. The physical seal is inserted through the locking clamps on the transmitter (if available on your transmitter).
Set up the Weights & Measures application 146 Micro Motion ® Model 2700 Transmitters with Analog Outputs.
Part III Operations, maintenance, and troubleshooting Chapters covered in this part: • Transmitter operation • Operate the transmitter with the Weights & Measures application • Measurement s.
Operations, maintenance, and troubleshooting 148 Micro Motion ® Model 2700 Transmitters with Analog Outputs.
9 Transmitter operation Topics covered in this chapter: • Record the process variables • View process variables • View transmitter status using the status LED • View and acknowledge status alarms • Read totalizer and inventory values • Start and stop totalizers and inventories • Reset totalizers • Reset inventories 9.
9.2 View process variables Display Scroll to the desired process variable. If AutoScroll is enabled, you can wait until the proc- ess variable is displayed.
Transmitter display features Figure 9-1: A B C D E F G H A. Status LED B. Display (LCD panel) C. Process variable D. Scroll optical switch E. Optical switch indicator: turns red when either Scroll or Select is activated F. Select optical switch G. Unit of measure for process variable H.
9.3 View transmitter status using the status LED The status LED shows the current alarm condition of the transmitter. The status LED is located on the face of the transmitter. Observe the status LED. • If your transmitter has a display, you can view the status LED with the transmitter housing cover in place.
Prerequisites Operator access to the alarm menu must be enabled (default setting). If operator access to the alarm menu is disabled, you must use another method to view or acknowledge status alarms.
Using the display to view and acknowledge the status alarms Figure 9-2: SEE ALARM Y es Scroll and Select simultaneously for 4 seconds ACK ALL Y es EXIT Select No Alarm code Scroll ACK Y es Select No A.
Postrequisites • To clear the following alarms, you must correct the problem, acknowledge the alarm, then power-cycle the transmitter: A001, A002, A010, A011, A012, A013, A018, A019, A022, A023, A024, A025, A028, A029, A031. • For all other alarms: - If the alarm is inactive when it is acknowledged, it will be removed from the list.
9.4.3 View and acknowledge alerts using ProLink III You can view a list containing all alerts that are active, or inactive and have been unacknowleged. From this list, you can acknowlege individual alerts or choose to acknowledge all alerts at once. 1.
• To refresh the list of active or unacknowledged alarms, press Service Tools > Alerts > Refresh Alerts . 9.4.5 Alarm data in transmitter memory The transmitter maintains three sets of data for every alarm that is posted.
Overview Totalizers keep track of the total amount of mass or volume measured by the transmitter since the last totalizer reset. Inventories keep track of the total amount of mass or volume measured by the transmitter since the last inventory reset.
1. Scroll until the word TOTAL appears in the lower left corner of the display. Important Because all totalizers are started or stopped together, it does not matter which total you use. 2. Select . 3. Scroll until START appears beneath the current totalizer value.
9.7 Reset totalizers Display See Section 9.7.1 . ProLink II ProLink > Totalizer Control > Reset Mass Total ProLink > Totalizer Control > Reset Volume Total ProLink > Totalizer Control &.
3. Scroll until RESET appears beneath the current totalizer value. 4. Select . 5. Select again to confirm. 6. Scroll to EXIT . 7. Select . • To reset the volume totalizer: 1. Scroll until the volume totalizer value appears. 2. Select . 3. Scroll until RESET appears beneath the current totalizer value.
Overview When you reset an inventory, the transmitter sets its value to 0. It does not matter whether the inventory is started or stopped. If the inventory is started, it continues to track process measurement. Tip When you reset a single inventory, the values of other inventories are not reset.
10 Operate the transmitter with the Weights & Measures application Topics covered in this chapter: • Operate the transmitter when the Weights & Measures application is installed • Switch b.
• You cannot stop totalizers while the tranmitter is secured. • You cannot reset inventories while the transmitter is secured. 10.1.1 Approved methods to read or obtain process data When the Weigh.
The display is able to show a maximum of eight characters, including the decimal point. For all totalizer values configured as display variables, the position of the decimal point on the display is fixed to the configured precision of the display variable.
Transmitter outputs and process data when Approvals= OIML Table 10-4: Function Transmitter status Unsecured Secured Outputs mA output behavior Performs configured fault action Normal Frequency output .
Available actions when Approvals= NTEP (continued) Table 10-5: Function Transmitter status Unsecured Secured Outputs set to Last Meas- ured Value Allowed Not allowed External pressure and tem- peratur.
Available actions when Approvals= OIML (continued) Table 10-6: Function Transmitter status Unsecured Secured Smart Meter Verification Outputs set to Continue Measuring Allowed Allowed Outputs set to F.
The controls are not accessible in any other way. 10.2.1 Switch between secured and unsecured mode using ProLink II Prerequisites Before switching to unsecured mode, ensure that you will be able to switch back to secured mode.
1. Make a service port connection to your transmitter. 2. Choose Device Tools > Configuration > Weights & Measures . 3. Set Software Security to Enabled . 10.2.3 Switch between secured and unsecured mode using the switching utility Prerequisites The switching utility must be installed on your PC.
10.4 Replacing the core processor in a Weights & Measures installation In a Weights & Measures installation, replacing the core processor requires re-sealing the transmitter. When the transmitter is first secured, the core processor’s unique ID is registered with the transmitter.
Operate the transmitter with the Weights & Measures application 172 Micro Motion ® Model 2700 Transmitters with Analog Outputs.
11 Measurement support Topics covered in this chapter: • Options for measurement support • Use Smart Meter Verification • Zero the flowmeter • Validate the meter • Perform a (standard) D1 and D2 density calibration • Perform a D3 and D4 density calibration (T-Series sensors only) • Perform temperature calibration 11.
11.2.1 Smart Meter Verification requirements To use Smart Meter Verification, the transmitter must be paired with an enhanced core processor, and the Smart Meter Verification option must be ordered for the transmitter.
Smart Meter Verification has an output mode called Continuous Measurement that allows the transmitter to keep measuring while the test is in progress. If you choose to run the test in Last Measured Value or Fault modes instead, the transmitter outputs will be held constant for the two minute duration of the test.
Option Description Last Value During the test, all outputs will go to their configured fault action. The test will run for approximately 140 seconds. While the test is in progress, dots traverse the display and test progress is shown. Postrequisites View the test results and take any appropriate actions.
Smart Meter Verification flowchart: Running a test using the display Running a Smart Meter Verification test using the display Figure 11-2: OUTPUTS ARE YOU SURE/YES? .
You may need to wait a few seconds while ProLink II synchronizes its database with the transmitter data. 2. Review the information presented on the screen, and click Next . 3. Enter any desired information on the Test Definition screen, and click Next .
Postrequisites View the test results and take any appropriate actions. Run a Smart Meter Verification test using the Field Communicator 1. Navigate to the Smart Meter Verification menu: • Overview > Shortcuts > Meter Verification • Service Tools > Maintenance > Routine Maintenance > Meter Verification 2.
• Current flowmeter identification data • Current flow and density configuration parameters • Current zero values • Current process values for mass flow rate, volume flow rate, density, temper.
Smart Meter Verification flowchart: Viewing test results using the display Viewing Smart Meter Verification test results using the display Figure 11-4: RESUL TS READ Select xx L STF% RUNCOUNT x Select.
View test result data using ProLink II 1. Choose Tools > Meter Verification > Run Meter Verification and click View Previous Test Results and Print Report . The chart shows test results for all tests stored in the ProLink II database. 2. (Optional) Click Next to view and print a test report.
Fail The test result is not within the specification uncertainty limit. Micro Motion recommends that you immediately repeat the meter verification test. If during the failed test you had set outputs to Continue Measurement , set outputs to Fault or Last Measured Value instead.
Manage scheduled test execution using the display 1. Navigate to the Smart Meter Verification menu. Smart Meter Verification – Top-level menu Figure 11-5: Scroll and Select simultaneously for 4 seconds ENTER METER VERFY Scroll RUN VERFY RESUL TS READ SCHEDULE VERFY Select EXIT Scroll Scroll Scroll Scroll Select Select Select Select 2.
Smart Meter Verification flowchart: Scheduling test execution using the display Scheduling Smart Meter Verification test execution using the display Figure 11-6: SCHEDULE VERFY Select SA VE/YES? TURN .
4. To disable scheduled execution: • To disable execution of a single scheduled test, set Hours Until Next Run to 0. • To disable recurring execution, set Hours Between Recurring Runs to 0. • To disable all scheduled execution, click Turn Off Schedule .
• The zero is required by site procedures. • The stored zero value fails the Zero Verification procedure. Prerequisites Before performing a field zero, execute the Zero Verification procedure to see whether or not a field zero can improve measurement accuracy.
2. Navigate to OFFLINE MAINT > ZERO > CAL ZERO and select CAL/YES? . Dots traverse the display while flowmeter zero is in progress. 3. Read the zero result on the display. The display reports CAL PASS if the zero was successful, or CAL FAIL if it was not.
3. Click Calibrate Zero . 4. Modify Zero Time , if desired. Zero Time controls the amount of time the transmitter takes to determine its zero- flow reference point. The default Zero Time is 20 seconds. For most applications, the default Zero Time is appropriate.
a. Allow the flowmeter to warm up for at least 20 minutes after applying power. b. Run the process fluid through the sensor until the sensor temperature reaches the normal process operating temperature. c. Stop flow through the sensor by shutting the downstream valve, and then the upstream valve if available.
11.3.4 Zero the flowmeter using the Field Communicator Zeroing the flowmeter establishes a baseline for process measurement by analyzing the sensor's output when there is no flow through the sensor tubes. 1. Prepare the flowmeter: a. Allow the flowmeter to warm up for at least 20 minutes after applying power.
11.4 Validate the meter Display OFF-LINE MAINT > CONFG > UNITS > MTR F ProLink II ProLink > Configuration > Flow ProLink III Device Tools > Configuration > Process Measurement >.
Procedure 1. Determine the meter factor as follows: a. Use the flowmeter to take a sample measurement. b. Measure the same sample using the reference device. c. Calculate the meter factor using the following formula: NewMeterFactor ConfiguredMeterFactor ReferenceMeasurement FlowmeterMeasurement = x 2.
MeterFactor V olume 1 MeterFactor Density = Note The following equation is mathematically equivalent to the first equation. You may use whichever version you prefer. MeterFactor V olume ConfiguredMeterFactor Density Density Flowmeter Density ReferenceDevice = x 3.
• If LD Optimization is enabled on your meter, disable it. To do this, choose ProLink > Configuration > Sensor and ensure that the checkbox is not checked. LD Optimizatio n is used only with large sensors in hydrocarbon applications. In some installations, only Micro Motion customer service has access to this parameter.
Postrequisites If you disabled LD Optimization before the calibration procedure, re-enable it. 11.5.2 Perform a D1 and D2 density calibration using ProLink III Prerequisites • During density calibra.
D1 and D2 density calibration using ProLink III Figure 11-8: Enter density of D1 fluid D1 calibration Close shutoff valve downstream from sensor Fill sensor with D1 fluid Done Device T ools > Calib.
• Before performing the calibration, record your current calibration parameters. If the calibration fails, restore the known values. Restriction For T-Series sensors, the D1 calibration must be performed on air and the D2 calibration must be performed on water.
11.6 Perform a D3 and D4 density calibration (T- Series sensors only) For T-Series sensors, the optional D3 and D4 calibration could improve the accuracy of the density measurement if the density of your process fluid is less than 0.8 g/cm 3 or greater than 1.
D3 or D3 and D4 density calibration using ProLink II Figure 11-10: Enter density of D3 fluid Calibration in Progress light turns green Calibration in Progress light turns red D3 calibration Close shut.
- Minimum difference of 0.1 g/cm 3 between the density of the D4 fluid and the density of the D3 fluid. The density of the D4 fluid must be greater than the density of the D3 fluid. - Minimum difference of 0.1 g/cm 3 between the density of the D4 fluid and the density of water.
• For D3 density calibration, the D3 fluid must meet the following requirements: - Minimum density of 0.6 g/cm 3 - Minimum difference of 0.1 g/cm 3 between the density of the D3 fluid and the density of water. The density of the D3 fluid may be either greater or less than the density of water.
D3 or D3 and D4 density calibration using the Field Communicator Figure 11-12: Enter density of D3 fluid Density Calibration Complete message Calibration in Progress message D3 calibration Close shuto.
Important Consult Micro Motion before performing a temperature calibration. Under normal circumstances, the temperature circuit is stable and should not need an adjustment.
Temperature calibration using ProLink III Figure 11-14: Enter temperature of low- temperature fluid T emperature Offset calibration W ait until sensor achieves thermal equilibrium Fill sensor with low.
Measurement support 206 Micro Motion ® Model 2700 Transmitters with Analog Outputs.
12 Troubleshooting Topics covered in this chapter: • Status LED states • Status alarms • Flow measurement problems • Density measurement problems • Temperature measurement problems • Milli.
12.1 Status LED states The status LED on the transmitter indicates whether or not alarms are active. If alarms are active, view the alarm list to identify the alarms, then take appropriate action to correct the alarm condition. Your transmitter has a status LED only if it has a display.
Status alarms and recommended actions (continued) Table 12-2: Alarm code Description Recommended actions A003 No Sensor Response The transmitter is not receiving one or more basic electrical sig- nals from the sensor.
Status alarms and recommended actions (continued) Table 12-2: Alarm code Description Recommended actions A005 Mass Flow Rate Overrange The sensor is signaling a flow rate that is out of range for the sen- sor. 1. If other alarms are present, resolve those alarm conditions first.
Status alarms and recommended actions (continued) Table 12-2: Alarm code Description Recommended actions A008 Density Overrange The sensor is signaling a density reading below 0 g/cm 3 or above 10 g/cm 3 .
Status alarms and recommended actions (continued) Table 12-2: Alarm code Description Recommended actions A010 Calibration Failure This alarm is typically caused by flow through the sensor during the zero, or by a zero offset result that is out of range.
Status alarms and recommended actions (continued) Table 12-2: Alarm code Description Recommended actions A016 Sensor RTD Failure The sensor RTD is signaling a resistance that is out of range for the sensor. 1. Check the wiring between the sensor and the transmitter.
Status alarms and recommended actions (continued) Table 12-2: Alarm code Description Recommended actions A019 RAM Error (Transmitter) Power to the transmitter must be cycled to clear this alarm. 1. Check that all wiring compartment covers are installed prop- erly.
Status alarms and recommended actions (continued) Table 12-2: Alarm code Description Recommended actions A026 Sensor/Transmitter Communi- cations Failure The transmitter has lost communication with the core processor on the sensor.
Status alarms and recommended actions (continued) Table 12-2: Alarm code Description Recommended actions A031 Low Power The core processor on the sensor is not receiving sufficient pow- er. Check the wiring between the transmitter and the sensor. Power to the transmitter must be cycled to clear this alarm.
Status alarms and recommended actions (continued) Table 12-2: Alarm code Description Recommended actions A100 mA Output 1 Saturated The calculated mA output value is outside of the meter's config- ured range. 1. Check the Upper Range Value and Lower Range Value parame- ters.
Status alarms and recommended actions (continued) Table 12-2: Alarm code Description Recommended actions A110 Frequency Output Saturated The calculated frequency output is outside the configured range. 1. Check the Frequency Output Scaling Method parameter.
Status alarms and recommended actions (continued) Table 12-2: Alarm code Description Recommended actions A117 Density Overrange (Petrole- um) 1. Check your process conditions against the values reported by the flowmeter. 2. Verify the configuration of the petroleum measurement ta- ble type and density.
12.3 Flow measurement problems Flow measurement problems and recommended actions Table 12-3: Problem Possible causes Recommended actions Flow indication at no flow conditions or zero offset • Misali.
Flow measurement problems and recommended actions (continued) Table 12-3: Problem Possible causes Recommended actions Erratic non-zero flow rate when flow is steady • Slug flow • Damping value too.
12.4 Density measurement problems Density measurement problems and recommended actions Table 12-4: Problem Possible causes Recommended actions Inaccurate density reading • Problem with process fluid.
12.5 Temperature measurement problems Temperature measurement problems and recommended actions Table 12-5: Problem Possible causes Recommended actions Temperature reading significantly different from process temper- ature • RTD failure • Wiring problem • Check junction box for moisture or verdi- gris.
12.6 Milliamp output problems Milliamp output problems and recommended actions Table 12-6: Problem Possible causes Recommended actions No mA output • Wiring problem • Circuit failure • Channel not configured for desired output • Check the power supply and power supply wiring.
Milliamp output problems and recommended actions (continued) Table 12-6: Problem Possible causes Recommended actions mA output consis- tently out of range • Incorrect process variable or units assig.
Frequency output problems and recommended actions Table 12-7: Problem Possible causes Recommended actions No frequency output • Stopped totalizer • Process condition below cutoff • Fault conditi.
For more information on using sensor simulation using ProLink II, see Section 7.1 . 12.9 Check power supply wiring If the power supply wiring is damaged or improperly connected, the transmitter may not receive enough power to operate properly. Prerequisites You will need the installation manual for your transmitter.
Prerequisites You will need the installation manual for your transmitter. Procedure 1. Before opening the wiring compartments, disconnect the power source. CAUTION! If the transmitter is in a hazardous area, wait five minutes after disconnecting the power.
Follow appropriate procedures to ensure that loop testing will not interfere with existing measurement and control loops. Procedure 1. Test the mA output(s). a. Choose and select a low value, e.g., 4 mA. Dots traverse the display while the output is fixed.
c. At the transmitter, activate Select . d. Scroll to and select SET OFF . e. Verify the signal at the receiving device. f. At the transmitter, activate Select . Postrequisites • If the mA output reading was slightly off at the receiving device, you can correct this discrepancy by trimming the output.
i. Click UnFix mA . 2. Test the frequency output(s). Note If the Weights & Measures application is enabled on the transmitter, it is not possible to perform a loop test of the frequency output, even when the transmitter is unsecured. a. Choose ProLink > Test > Fix Freq Out .
Procedure 1. Test the mA output(s). a. Choose Device Tools > Diagnostics > Testing > mA Output 1 Test or Device Tools > Diagnostics > Testing > mA Output 2 Test . b. Enter 4 in Fix to: . c. Click Fix mA . d. Read the mA current at the receiving device and compare it to the transmitter output.
f. Click UnFix . 12.12.4 Perform loop tests using the Field Communicator Tip Loop tests are not required. However, Micro Motion recommends performing a loop test for every input or output available on your transmitter. The inputs and outputs available on your transmitter vary according to your purchase option and your channel configuration.
b. Read the frequency signal at the receiving device and compare it to the transmitter output. c. Choose End . 3. Test the discrete output(s). a. Press Service Tools > Simulate > Simulate Outputs > Discrete Output Test . b. Choose Off . c. Verify the signal at the receiving device.
12.14 Check the HART communication loop If you cannot establish or maintain HART communications, the HART loop may be wired incorrectly. Prerequisites You will need: • A copy of your transmitter installation manual • A Field Communicator • Optional: the HART Application Guide , available at www.
Procedure 1. Set HART Address as appropriate for your HART network. The default address is 0. This is the recommended value unless the transmitter is in a multidrop network. 2. Set Loop Current Mode to Enabled . 12.16 Check HART burst mode HART burst mode can cause the transmitter to output unexpected values.
• For the relevant status alarms, change the setting of Alarm Severity to Ignore . 3. If there are no active fault conditions, continue troubleshooting. 12.20 Check Frequency Output Maximum Pulse Width If Frequency Output Maximum Pulse Width is set incorrectly, the frequency output may report an incorrect value.
The Flow Direction parameter interacts with actual flow direction to affect flow values, flow totals and inventories, and output behavior. For the simplest operation, actual process flow should match the flow arrow that is on the side of the sensor case.
12.26 Check the drive gain Excessive or erratic drive gain may indicate any of a variety of process conditions, sensor problems, or configuration problems. To know whether your drive gain is excessive or erratic, you must collect drive gain data during the problem condition and compare it to drive gain data from a period of normal operation.
Possible causes and recommended actions for erratic drive gain (continued) Table 12-9: Possible cause Recommended actions Polarity of pick-off reversed or polarity of drive reversed Contact Micro Motion. Slug flow Check for slug flow. See Section 12.25 .
Possible causes and recommended actions for low pickoff voltage (continued) Table 12-10: Possible cause Recommended actions No tube vibration in sensor • Check for plugging. • Ensure sensor is free to vibrate (no mechanical binding). • Verify wiring.
12.28.1 Check the sensor coils Checking the sensor coils can identify electrical shorts. Restriction This procedure applies only to 9-wire remote-mount transmitters and remote transmitters with remote core processors.. Procedure 1. Disconnect power to the transmitter.
a. Leave the terminal blocks disconnected. b. Remove the lid of the junction box. c. Testing one terminal at a time, place a DMM lead on the terminal and the other lead on the sensor case. With the DMM set to its highest range, there should be infinite resistance on each lead.
a. Remove the core processor lid. The core processor is intrinsically safe and can be opened in all environments. b. Check the state of the core processor LED. 3. If you have an integral installation: a. Loosen the four cap screws that fasten the transmitter to the base.
9-wire remote installation components Figure 12-2: Transmitter Core processor 4 x cap screws (4 mm) End-cap b. Inside the core processor housing, loosen the three screws that hold the core processor mounting plate in place. Do not remove the screws. c.
3. Tighten the screws, torquing to 6 to 8 in-lbs (0.7 to 0.9 N-m). 4. Replace the end-cap. Important When reassembling the meter components, be sure to grease all O-rings.
Enhanced core processor LED states (continued) Table 12-14: LED state Description Recommended action Solid red High-severity alarm Check alarm status. Flashing red (80% on, 20% off) Tubes not full • If alarm A105 (slug flow) is active, refer to the recommended actions for that alarm.
Integral installation components Figure 12-3: Base 4 x cap screws (4 mm) Transition ring Transmitter Core processor b. Rotate the transmitter counter-clockwise so that the cap screws are in the unlocked position. c. Gently lift the transmitter straight up, disengaging it from the cap screws.
5. At the core processor, disconnect the 4-wire cable between the core processor and the transmitter. 6. Measure the resistance between core processor terminal pairs 3–4, 2–3, and 2–4.
Troubleshooting 250 Micro Motion ® Model 2700 Transmitters with Analog Outputs.
Appendix A Using the transmitter display Topics covered in this appendix: • Components of the transmitter interface • Use the optical switches • Access and use the display menu system • Display codes for process variables • Codes and abbreviations used in display menus • Menu maps for the transmitter display A.
Transmitter interface Figure A-1: A B C D E F G H A. Status LED B. Display (LCD panel) C. Process variable D. Scroll optical switch E. Optical switch indicator F. Select optical switch G. Unit of measure for process variable H. Current value of process variable A.
Optical switch indicator and optical switch states Table A-1: Optical switch indicator State of optical switches Solid red One optical switch is activated.
Tip If you do not know the correct value for Off-Line Password , wait 30 seconds. The password screen will time out automatically and you will be returned to the previous screen. 4. If Scroll flashes on the display, activate the Scroll optical switch, then the Select optical switch, and then the Scroll optical switch again.
Procedure • To change the value: 1. Activate Select until the digit you want to change is active (flashing). Select moves the cursor one position to the left. From the leftmost position, Select moves the cursor to the rightmost digit. 2. Activate Scroll to change the value of the active digit.
• To exit the menu without saving the displayed value to transmitter memory, activate Scroll and Select simultaneously and hold until the display changes. - If the displayed value is the same as the value in transmitter memory, you will be returned to the previous screen.
b. Activate Scroll until the desired character is displayed. c. Activate Select to move the cursor one digit to the left. d. Activate Scroll until the desired character is displayed. e. Activate Select to move the cursor one digit to the left. f. Activate Scroll until the desired character is displayed.
Display codes for process variables (continued) Table A-2: Code Definition Comment or reference FVZ Field verification zero Weights & Measures application only GSV F Gas standard volume flow GSV I.
A.5 Codes and abbreviations used in display menus Codes and abbreviations used in display menus Table A-3: Code or abbrevi- ation Definition Comment or reference ACK ALARM Acknowledge alarm ACK ALL Ac.
Codes and abbreviations used in display menus (continued) Table A-3: Code or abbrevi- ation Definition Comment or reference ENABLE ALARM Enable alarm menu Access to alarm menu from display ENABLE AUTO.
Codes and abbreviations used in display menus (continued) Table A-3: Code or abbrevi- ation Definition Comment or reference MASS Mass flow MBUS Modbus MFLOW Mass flow MSMT Measurement OFFLN Off-line OFF-LINE MAINT Off-line maintenance P/UNT Pulses/unit POLAR Polarity PRESS Pressure QUAD Quadrature r.
A.6 Menu maps for the transmitter display Offline menu – top level Figure A-2: Scroll and Select simultaneously for 4 seconds VER OFF-LINE MAINT Select SEE ALARM Scroll Scroll EXIT CONFG Scroll SIM .
Offline menu – version information Figure A-3: Scroll and Select simultaneously for 4 seconds VER Y es V ersion info Scroll Select Y es CEQ/ETO info* Scroll CUSTODY XFER* Scroll Y es SENSOR VERFY* S.
Offline menu – configuration: units and I/O Figure A-4: OFF-LINE MAINT Scroll and Select simultaneously for 4 seconds Scroll Select Select Scroll CONFG MASS UNITS VOL/GSV DENS TEMP Select Scroll Scr.
Offline menu – configuration: meter factor, display, and digital communications Figure A-5: OFF-LINE MAINT Scroll and Select simultaneously for 4 seconds Scroll Select Select Scroll CONFG UNITS MASS.
Offline menu – alarms Figure A-6: SEE ALARM Scroll and Select simultaneously for 4 seconds ACK ALL* Y es EXIT Select No Alarm code Scroll ACK Y es Select No Active/ unacknowledged alarms? No Y es Se.
Offline menu – meter verification: top level Figure A-7: Scroll and Select simultaneously for 4 seconds ENTER METER VERFY Scroll RUN VERFY RESUL TS READ SCHEDULE VERFY Select EXIT Scroll Scroll Scro.
Offline menu – meter verification test Figure A-9: OUTPUTS ARE YOU SURE/YES? . . . . . . . . . . . . . . . x % P ASS VERFY ABORTED VERFY CAUTION VERFY Fail Abort RERUN/YES? Y es No Correct condition.
Offline menu – meter verification results Figure A-10: RESUL TS READ Select xx L STF% RUNCOUNT x Select xx HOURS Select P ASS Select xx R STF% Select RESUL TS MORE? Select Scroll T o Run V erfy Pass.
Offline menu – totalizers and inventories Figure A-11: RESET (3) Select Scroll STOP/ST ART (2) RESET YES? Process variable display STOP/ST ART YES? Scroll Mass total V olume total Scroll Select Y es.
Offline menu – Simulation (loop testing) Figure A-12: Scroll and Select simultaneously for 4 seconds Y es Scroll Select AO SIM FO SIM DO SIM Scroll Select SET x MA* Y es Select** SET y KHZ**** Select SET OFF SET ON Select EXIT Scroll Scroll Select*** .
Offline menu – Zero Figure A-13: …………………. OFF-LINE MAINT Scroll and Select simultaneously for 4 seconds Scroll Select Select CAL ZERO Troubleshoot ZERO/YES? CAL P ASS CAL F AIL RESTORE.
Appendix B Using ProLink II with the transmitter Topics covered in this appendix: • Basic information about ProLink II • Connect with ProLink II • Menu maps for ProLink II B.1 Basic information about ProLink II ProLink II is a software tool available from Micro Motion.
ProLink II messages As you use ProLink II with a Micro Motion transmitter, you will see a number of messages and notes. This manual does not document all of these messages and notes. Important The user is responsible for responding to messages and notes and complying with all safety messages.
B.2.2 Make a service port connection CAUTION! If the transmitter is in a hazardous area, do not use a service port connection. Service port connections require opening the wiring compartment, and opening the wiring compartment while the transmitter is powered up could cause an explosion.
Connection to service port Figure B-1: A C D E B A. PC B. Signal converter C. Service port terminal 7 (RS-485/A) D. Service port terminal 8 (RS-485/B) E. Transmitter, with wiring compartment and power supply compartment opened Note This figure shows a serial port connection.
CAUTION! If the transmitter is in a hazardous area, do not connect directly to the transmitter terminals. Connecting directly to the transmitter terminals requires opening the wiring compartment, and opening the wiring compartment while the transmitter is powered up could cause an explosion.
Connection to transmitter terminals Figure B-2: A C D B A. PC B. Signal converter C. 250–600 Ω resistance D. Transmitter, with wiring compartment and power supply compartment opened Note This figure shows a serial port connection. USB connections are also supported.
Connection over local loop Figure B-3: A C D E R1 R3 R2 B A. PC B. Signal converter C. Any combination of resistors R1, R2, and R3 as necessary to meet HART communication resistance requirements D. DCS or PLC E. Transmitter, with wiring compartment and power supply compartment opened Note This figure shows a serial port connection.
Connection over multidrop network Figure B-4: B A C D A. Signal converter B. 250–60 0 Ω resistance C. Devices on the network D. Master device 5. Start ProLink II. 6. Choose Connection > Connect to Device . 7. Set Protocol to HART Bell 202 . Tip HART/Bell 202 connections use standard connection parameters.
Option Description Primary Use this setting if no other host is on the network. The Field Communicator is not a host. 12. Click Connect . Need help? If an error message appears: • Verify the HART address of the transmitter. • Ensure that you have specified the correct COM port.
Tip HART connections are not polarity-sensitive. It does not matter which lead you attach to which terminal. Connection to transmitter terminals Figure B-5: A C B A. PC B. Signal converter C. Transmitter, with wiring compartment and power supply compartment opened Note This figure shows a serial port connection.
Connection over network Figure B-6: A C E D B A. PC B. Signal converter C. 120- Ω , 1/2-watt resistors at both ends of the segment, if necessary D. DCS or PLC E. Transmitter, with wiring compartment and power supply compartment opened Note This figure shows a serial port connection.
Option Description Secondary Use this setting if another HART host such as a DCS is on the network. Primary Use this setting if no other host is on the network. The Field Communicator is not a host. 9. Click Connect . Need help? If an error message appears: • Verify the HART address of the transmitter.
Tip Usually, but not always, the black lead is RS-485/A and the red lead is RS-485/B. Connection to transmitter terminals Figure B-7: A C B A. PC B. Signal converter C. Transmitter, with wiring compartment and power supply compartment opened Note This figure shows a serial port connection.
Connection over network Figure B-8: A C E D B A. PC B. Signal converter C. 120- Ω , 1/2-watt resistors at both ends of the segment, if necessary D. DCS or PLC E. Transmitter, with wiring compartment and power supply compartment opened Note This figure shows a serial port connection.
Need help? If an error message appears: • Verify the Modbus address of the transmitter. • Ensure that you have specified the correct COM port. • Check the physical connection between the PC and the transmitter. • Increase or decrease resistance.
Main menu (continued) Figure B-10: Data Logging* Enable/Disable Custody T ransfer T ools Plug-ins ProLink Configuration Output Levels Process V ariables Status Alarm Log Diagnostic Information Calibra.
Configuration menu Figure B-11: Flow • Flow Direction • Flow Damp • Flow Cal • Mass Flow Cutoff • Mass Flow units • Mass Factor • Dens Factor • V ol Factor • Flow Switch V ariable .
Configuration menu (continued) Figure B-12: T emperature • T emp Units • T emp Cal Factor • T emp Damping • External T emperature • External RTD Frequency/Discrete Output • Frequency • T.
Configuration menu (continued) Figure B-13: ProLink > Configuration Additional configuration options Device • Model • Manufacturer • Hardware Rev • Distributor • Software Rev • ET O •.
Configuration menu (continued) Figure B-14: ProLink > Configuration Polled V ariables Pol led V ariable 1/2 • Polling Control • External T ag • V ariable T ype • Current V alue Discrete Eve.
Configuration menu (continued) Figure B-15: ProLink > Configuration Additional configuration options Display • mA1 • V ar1...V ar15 • Display Precision • V ar • Number of Decimals • Dis.
Configuration menu (continued) Figure B-16: ProLink > Configuration T ransmitter Options • V olume Flow • Meter Fingerprinting • Cryogenic Moduls Compensation • Core Processor Options • B.
Configuration menu (continued) Figure B-17: ProLink > Configuration Sensor • Sensor s /n • Sensor Model • Sensor Matl • Liner Matl • Flange API Setup • Generalized Crude or JP 4 • Gen.
Configuration menu (continued) Figure B-18: ProLink > Configuration Sensor Simulation • Enable/disable • Mass flow • W ave form • Fixed value • Period • Minimum • Maximum • Density .
Appendix C Using ProLink III with the transmitter Topics covered in this appendix: • Basic information about ProLink III • Connect with ProLink III • Menu maps for ProLink III C.1 Basic information about ProLink III ProLink III is a configuration and service tool available from Micro Motion.
ProLink III messages As you use ProLink III with a Micro Motion transmitter, you will see a number of messages and notes. This manual does not document all of these messages and notes. Important The user is responsible for responding to messages and notes and complying with all safety messages.
C.2.2 Make a service port connection CAUTION! If the transmitter is in a hazardous area, do not use a service port connection. Service port connections require opening the wiring compartment, and opening the wiring compartment while the transmitter is powered up could cause an explosion.
Connection to service port Figure C-1: A C D E B A. PC B. Signal converter C. Service port terminal 7 (RS-485/A) D. Service port terminal 8 (RS-485/B) E. Transmitter, with wiring compartment and power supply compartment opened Note This figure shows a serial port connection.
CAUTION! If the transmitter is in a hazardous area, do not connect directly to the transmitter terminals. Connecting directly to the transmitter terminals requires opening the wiring compartment, and opening the wiring compartment while the transmitter is powered up could cause an explosion.
Connection to transmitter terminals Figure C-2: A C D B A. PC B. Signal converter C. 250–600 Ω resistance D. Transmitter, with wiring compartment and power supply compartment opened Note This figure shows a serial port connection. USB connections are also supported.
Connection over local loop Figure C-3: A C D E R1 R3 R2 B A. PC B. Signal converter C. Any combination of resistors R1, R2, and R3 as necessary to meet HART communication resistance requirements D. DCS or PLC E. Transmitter, with wiring compartment and power supply compartment opened Note This figure shows a serial port connection.
Connection over multidrop network Figure C-4: B A C D A. Signal converter B. 250–60 0 Ω resistance C. Devices on the network D. Master device 5. Start ProLink III. 6. Choose Connect to Physical Device . 7. Set Protocol to HART Bell 202 . Tip HART/Bell 202 connections use standard connection parameters.
Option Description Primary Use this setting if no other host is on the network. The Field Communicator is not a host. 12. Click Connect . Need help? If an error message appears: • Verify the HART address of the transmitter. • Ensure that you have specified the correct COM port.
Tip HART connections are not polarity-sensitive. It does not matter which lead you attach to which terminal. Connection to transmitter terminals Figure C-5: A C B A. PC B. Signal converter C. Transmitter, with wiring compartment and power supply compartment opened Note This figure shows a serial port connection.
Connection over network Figure C-6: A C E D B A. PC B. Adapter, if necessary C. Signal converter D. 120- Ω , 1/2-watt resistors at both ends of the segment, if necessary E. DCS or PLC F. Transmitter, with wiring compartment and power supply compartment opened Note This figure shows a serial port connection.
8. Set Master as appropriate. Option Description Secondary Use this setting if another HART host such as a DCS is on the network. Primary Use this setting if no other host is on the network. The Field Communicator is not a host. 9. Click Connect . Need help? If an error message appears: • Verify the HART address of the transmitter.
b. Connect the leads from the signal converter to terminals 5 (RS-485/A) and 6 (RS-485/B). Tip Usually, but not always, the black lead is RS-485/A and the red lead is RS-485/B. Connection to transmitter terminals Figure C-7: A C B A. PC B. Signal converter C.
Connection over network Figure C-8: A C E D B A. PC B. Signal converter C. 120- Ω , 1/2-watt resistors at both ends of the segment, if necessary D. DCS or PLC E. Transmitter, with wiring compartment and power supply compartment opened Note This figure shows a serial port connection.
Need help? If an error message appears: • Verify the Modbus address of the transmitter. • Ensure that you have specified the correct COM port. • Check the physical connection between the PC and the transmitter. • Increase or decrease resistance.
Device Tools: Configuration (with Weights & Measures) Figure C-10: Configuration: Process Measurement (with Concentration Measurement) Figure C-11: Using ProLink III with the transmitter 312 Micro.
Configuration: Process Measurement (with Petroleum Measurement) Figure C-12: Configuration: I/O Figure C-13: Using ProLink III with the transmitter Configuration and Use Manual 313.
Configuration: Transmitter Display Figure C-14: Configuration: Events Figure C-15: Using ProLink III with the transmitter 314 Micro Motion ® Model 2700 Transmitters with Analog Outputs.
Configuration: Communications Figure C-16: Configuration: Informational Parameters Figure C-17: Using ProLink III with the transmitter Configuration and Use Manual 315.
Device Tools: Calibration Figure C-18: Calibration: Density Calibration Figure C-19: Using ProLink III with the transmitter 316 Micro Motion ® Model 2700 Transmitters with Analog Outputs.
Calibration: Temperature Calibration Figure C-20: Device Tools: Configuration Transfer Figure C-21: Using ProLink III with the transmitter Configuration and Use Manual 317.
Diagnostics: Testing Figure C-22: Diagnostics: Meter Verification Figure C-23: Using ProLink III with the transmitter 318 Micro Motion ® Model 2700 Transmitters with Analog Outputs.
Device Tools: Trending Figure C-24: Using ProLink III with the transmitter Configuration and Use Manual 319.
Using ProLink III with the transmitter 320 Micro Motion ® Model 2700 Transmitters with Analog Outputs.
Appendix D Using the Field Communicator with the transmitter Topics covered in this appendix: • Basic information about the Field Communicator • Connect with the Field Communicator • Menu maps for the Field Communicator D.
If Micro Motion is not listed, or you do not see the required device description, use the Field Communicator Easy Upgrade Utility to install the device description, or contact Micro Motion. Field Communicator menus and messages Many of the menus in this manual start with the On-Line menu.
Field Communicator connection to transmitter terminals Figure D-1: A B C A. Field Communicator B. 250–60 0 Ω resistance C. Transmitter, with wiring compartment and power supply compartment opened 2.
Field Communicator connection to multidrop network Figure D-3: A B C D A. Field Communicator B. 250–60 0 Ω resistance C. Devices on the network D. Master device 4. Turn on the Field Communicator and wait until the main menu is displayed. 5. If you are connecting across a multidrop network: a.
On-Line menu Figure D-4: Configure 1 Manual Setup 2 Alert Setup Service T ools 1 Alerts 2 V ariables 3 T rends 4 Maintenance 5 Simulate 2 3 Overview 1 Check Status 2 Primary Purpose V ariables 3 Short.
Overview menu Figure D-5: Identification 1 T ag 2 Model 3 Xmtr Serial Num 4 Sensor Serial Num 5 Date 6 Descriptor 7 Message 1 Revisions 1 Universal 2 Field Device 3 DD Revision 4 T ransmitter Software 5 CP Software 6 ETO Number Mat. of Construction 1 T ube Wetted Mat.
Configure menu Figure D-6: Manual Setup 1 Characterize 2 Measurements 3 Display 4 Inputs/Outputs 5 Info Parameters 2 1 Alert Setup 1 I/O Fault Actions 2 Alert Severity 3 Discrete Events 4 CM Alerts On.
Manual Setup menu Figure D-7: Characterize 1 Sensor T ype 2 Sensor T ag Parameters Measurements 1 Flow 2 Density 3 T emperature 4 Update Rate 5 LD Optimization 6 Special Units 7 External Pressure/T em.
Manual Setup menu: Characterize Figure D-8: On-Line Menu > 2 Configure > 1 Manual Setup > 1 Characterize 2 1 2 1 Sensor T ype Straight T ube Curved T ube Density Parameters 1 D1 2 D2 3 DT 4 D.
Manual Setup menu: Measurements Figure D-9: On-Line Menu > 2 Configure > 1 Manual Setup > 2 Measurements Flow 1 Flow Direction 2 Flow Damping 3 Mass Flow Unit 4 Mass Flow Cutoff 5 V olume Flo.
Manual Setup menu: Measurements (continued) Figure D-10: On-Line Menu > 2 Configure > 1 Manual Setup > 2 Measurements 1 2 3 8 7 9 10 Pressure 1 Pressure Unit 2 Pressure Compensation 3 Compens.
Manual Setup menu: Display Figure D-11: On-Line Menu > 2 Configure > 1 Manual Setup > 3 Display Language English German French Spanish Display V ariable Menu Features 1 T otalizer Reset 2 Sta.
Manual Setup menu: I/O Figure D-12: On-Line Menu > 2 Configure > 1 Manual Setup > 4 Inputs/Outputs mA Output 1 Primary V ariable 2 mA Output Settings 3 mA Fault Settings Frequency Output 1 FO.
Manual Setup menu: I/O (continued) Figure D-13: Discrete Output 1 DO Assignment 2 DO Polarity 3 DO Fault Action 4 Flow Switch Source 5 Flow Switch Setpoint 6 Hysteresis (0.
Alert Setup menu Figure D-14: On-Line Menu > 2 Configure > 2 Alert Setup I/O Fault Action 1 MAO Fault Action 2 MAO Fault Level 3 FO Fault Action 4 FO Fault Level 5 Comm Fault Action Alert Severi.
Service Tools menu Figure D-15: On-Line Menu > 3 Service T ools Alerts 1 Refresh Alerts Alert Name Additional Information V ariables 1 V ariable Summary 2 Process V ariables 3 Mapped V ariables 4 E.
Service Tools menu: Variables Figure D-16: On-Line Menu > 3 Service T ools > 2 V ariables Process V ariables 1 Mass Flow Rate 2 V olume Flow Rate * 3 Density 4 T emperature Mapped V ariables 1 P.
Service Tools menu: Variables (continued) Figure D-17: Outputs 1 mA Output (MA0) 2 Frequency Output On-Line Menu > 3 Service T ools > 2 V ariables V ariables (PM) * 1 Density at Reference T empe.
Service Tools menu: Maintenance Figure D-18: On-Line Menu > 3 Service T ools > 4 Maintenance Routine Maintenance 1 T rim mA Output 2 Meter V erification * 1 Zero Calibration 1 Mass Flow Rate 2 V.
Service Tools menu: Simulate Figure D-19: On-Line Menu > 3 Service T ools > 5 Simulate 1 Simulate Outputs 1 mA Output Loop T est 2 Frequency Output T est/ Discrete Output T est * * Options vary depending on Channel settings .
Appendix E Default values and ranges E.1 Default values and ranges The default values and ranges represent the typical factory transmitter configuration. Depending on how the transmitter was ordered, certain values may have been configured at the factory and are not represented in the default values and ranges.
Transmitter default values and ranges (continued) Table E-1: Type Parameter Default Range Comments Density units g/cm 3 Density cutoff 0.2 g/cm 3 0.0 – 0.5 g/cm 3 D1 0 g/cm 3 D2 1 g/cm 3 K1 1000 µsec 1000 – 50,000 µsec K2 50,000 µsec 1000 – 50,000 µsec FD 0 Temp Coefficient 4.
Transmitter default values and ranges (continued) Table E-1: Type Parameter Default Range Comments Base volume time sec Volume flow conversion factor 1 Variable map- ping Primary variable Mass flow Secondary variable Density Tertiary variable Mass flow Quaternary variable Volume flow mA output 1 Primary variable Mass flow LRV –200.
Transmitter default values and ranges (continued) Table E-1: Type Parameter Default Range Comments Fault action Downscale AO fault level – downscale 2.0 mA 1.0 – 3.6 mA AO fault level – upscale 22 mA 21.0 – 24.0 mA Last measured value timeout 0.
Transmitter default values and ranges (continued) Table E-1: Type Parameter Default Range Comments Polarity Active low Display Backlight on/off On Backlight intensity 63 0 – 63 Refresh rate 200 mill.
Default values and ranges 346 Micro Motion ® Model 2700 Transmitters with Analog Outputs.
Appendix F Transmitter components and installation wiring Topics covered in this appendix: • Installation types • Power supply terminals and ground • Input/output (I/O) wiring terminals F.1 Installation types Model 1700 and Model 2700 transmitters can be installed five different ways, only one of which applies to your specific installation.
High-temperature flexible conduit installation Figure F-2: High-temperature flexible conduit installations use the same installation instructions as 4-wire remote installations, except that the distance between the sensor and the electronics is limited by the length of the flexible conduit.
4-wire remote installation – stainless steel housing Figure F-4: Sensor Core processor Transmitter 4-wire cable • 9-wire remote – The transmitter and core processor are combined in a single unit that is installed remotely from the sensor.
9-wire remote installation type Figure F-5: Transmitter Junction box Sensor 9-wire cable • Remote core processor with remote sensor – A remote core process with remote sensor installation separates all three components – transmitter, core processor, and sensor – all of which are installed separately.
Remote core processor with remote sensor installation type Figure F-6: Core processor Transmitter 4-wire cable 9-wire cable Sensor Junction box F.2 Power supply terminals and ground Power supply wiring terminals Figure F-7: A B C A. Warning flap B. Equipment ground C.
F.3 Input/output (I/O) wiring terminals I/O wiring terminals Figure F-8: A B C A. mA/HART B. Frequency output or discrete output C. RS-485 Transmitter components and installation wiring 352 Micro Moti.
Appendix G NE 53 history G.1 NE 53 history Date Version Type Change Operating in- struction 08/2000 1.x Expansion Added writing of the device tag using Modbus 3600204 A Adjustment Improved communication handling with the HART Tri-Loop product Feature Indication of outputs option board type appears on display at power-up 05/2001 2.
Date Version Type Change Operating in- struction The display start/stop totalizers function can be enabled or disabled Petroleum measurement application improve- ments Live zero available as display v.
Date Version Type Change Operating in- struction 09/2006 5.x Expansion Discrete output assignable as a flow switch 20001715 B Discrete output fault indication configurability Discrete input support fo.
Date Version Type Change Operating in- struction Adjustment The following combinations are not allowed: • mA Output Fault Action = None and Digital Communications Fault Action = NAN • Frequency Ou.
Index A Added Damping 105 Additional Communications Response Delay 130 address HART address 124, 129 Modbus address 130 air calibration , See calibration, density alarm menu , See display alarms alarm.
restore factory configuration using ProLink II 22 using ProLink III 22 temperature measurement 50 volume flow measurement 29 Weights & Measures , See Weights & Measures application write-prote.
polarity configuring 118 options 118 source configuring 115 options 115 display accessing menu system 253 components 251 configuring display behavior Auto Scroll 81 backlight 81 display language 77 di.
using the display 228 using the Field Communicator 233 maximum pulse width 112 polarity configuring 110 options 110 process variable configuring 108 options 109 scaling method configuring 111 Frequenc.
process variable configuring 100 options 101 scaling 102 troubleshooting 224, 236 mass flow measurement configuring 23 cutoff configuring 27 effect on volume measurement 28 interaction with AO cutoff .
using ProLink III 54 using the Field Communicator 56 polling address , See HART address power power up 7 power supply terminals 351 power supply wiring troubleshooting 227 pressure compensation config.
slave address , See Modbus address slug flow , See density measurement, slug flow Smart Meter Verification automatic execution and scheduling using ProLink II 185 using ProLink III 186 using the displ.
unsecured , See Weights & Measures application Update Period 80 Update Rate configuring 87 effect on process measurement 88 incompatible functions 88 Upper Range Value (URV) 102 V Volume Correctio.
.
*MMI-20019043* MMI-20019043 Rev AA 2012 Micro Motion Inc. USA Worldwide Headquarters 7070 Winchester Circle Boulder, Colorado 80301 T +1 303-527-5200 T +1 800-522-6277 F +1 303-530-8459 www.micromotion.com Micro Motion Europe Emerson Process Management Neonstraat 1 6718 WX Ede The Netherlands T +31 (0) 318 495 555 F +31 (0) 318 495 556 www.
An important point after buying a device Emerson MMI-20019043 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Emerson MMI-20019043 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Emerson MMI-20019043 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Emerson MMI-20019043 you will learn all the available features of the product, as well as information on its operation. The information that you get Emerson MMI-20019043 will certainly help you make a decision on the purchase.
If you already are a holder of Emerson MMI-20019043, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Emerson MMI-20019043.
However, one of the most important roles played by the user manual is to help in solving problems with Emerson MMI-20019043. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Emerson MMI-20019043 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center