Carbon Dioxide Sensor Product Information Sheet 210-2000 Overview CPC’s CO 2 sensor (P/N 210 -2000) is a non-dispers ive infrared analyzer designed for measuring environm ental CO 2 conc entratio n in ventila tion systems and indoor spaces.
Carbon Dioxide Sensor Product Information Sheet 210-2000 Wiring The mounting plate has two sets of th ree-terminal connectors. T he connecto r on the left side of th e mounting plate is wher e sensor power and the signa l- to-I/O-board connection is made.
Carbon Dioxide Sensor Product Information Sheet 210-2000 Setting the Volts/Amps Switch The front of the circu it board has a switch that controls whether the analog outpu t sends a signal as voltage (V) or mi lliamps (mA). To work with CPC I/ O boards, this switch should be set to voltage (the DOWN position).
Carbon Dioxide Sensor Product Information Sheet 210-2000 Span The Span parameter determines whether you want to measure concen trations between 0-2000 ppm or 0 - 5000 ppm. If set to 2000 ppm, the sen sor’s analog output signal will be scaled proper ly so that the highest output voltag e corresponds to 2000 ppm.
An important point after buying a device Emerson 210-2000 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Emerson 210-2000 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Emerson 210-2000 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Emerson 210-2000 you will learn all the available features of the product, as well as information on its operation. The information that you get Emerson 210-2000 will certainly help you make a decision on the purchase.
If you already are a holder of Emerson 210-2000, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Emerson 210-2000.
However, one of the most important roles played by the user manual is to help in solving problems with Emerson 210-2000. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Emerson 210-2000 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center