Agilent Specifying Calibration Standards for the Agilent 8510 Network Analyzer Product Note 8510-5B.
2 3 3 3 4 5 5 7 7 8 9 9 9 11 12 12 12 14 14 15 16 16 17 17 17 18 18 18 18 19 19 19 19 19 20 20 20 21 21 22 22 22 23 23 26 29 Introduction Measurement errors Measurement calibration Calibration kit Sta.
3 Introduction This pr oduct note co vers measurement calibration requirements f or the Agilent 8510B/C network analyzer . All of t he capabilities described in this note also apply to the A gilent 85.
4 The arra y coeff icients are computed by measuring a set of “kno wn” devices connected at a fixed point and sol ving as the vector difference between the modeled and measured response. The full 2-por t er r or model sho wn in Figure 1 is an exam ple of only one of the measurement calibra- tions a vailable with the 8510.
5 Standard definition Standard definition is t he process of mathematical- ly modeling the electrical charact eristics (delay , attenuation and impedance) of each calibration standard.
6 T able 1. Standard definitions table T able 2. Standard class assignments.
7 Modification procedure Calibration kit modif ication pro vides the capability to adapt t o measurement calibrations in other con- nector types or to generate more precise error models fr om existing kits.
8 Define standards A glossar y of standard definition parameters used with the Agilent 8510 is included in this section. Each parameter is described and appropriate con- versions are listed f or im plement ation with t he 8510. T o illustrate, a calibration kit f or WR-62 rec- tangular wav eguide (operating frequency range 12.
9 Each standard is described using the Standard Definition Table in accordance wit h t he 1- or 2- port model. The St andard Definition t able for a wav eguide calibration kit is sho wn in Table 1. Each standard type (shor t, open, load, thru, and arbi- trar y impedance) ma y be def ined by the parame- ters as specif ied below .
10 It is not possible t o remov e fringing capacit ance, but the result ant phase shif t can be modeled as a function of frequency using C 0 t hr ough C 3 (C 0 +C l f + C 2 f 2 + C 3 f 3 ,.
11 Note In some cases (when t he phase response is linear with respect t o frequency) the response of an open can be modeled as an equivalent “increment al” length. This method will serve as a f irst order appro xima- tion onl y , but can be useful when data or st an- dards for t he abov e modeling techniques are not available.
The inductance as a function of frequency can be modeled b y specifying t he coef ficients of a third- order polynomial (L 0 + L 1 f + L 2 f 2 + L 3 f 3 ), with units of L 0 (nH), L 1 (10 -24 H/Hz), L 2 (10 -33 H/Hz 2 ) and L 3 (10 -42 H/Hz 3 ).
13 The conv ention f or definition of offset dela y in wav eguide requires entr y of t he delay assuming no dispersion. F or wa veguide transmission line, the Agilent 8510 calculat es t he ef f ects o.
14 µ r = relativ e permeability constant of t he medium (equal to 1.0 in air) ε r = relativ e permittivity constant of t he medium (equal to 1.000649 in air) D = inside diameter of outer conductor d.
15 Therefore, for the WR -62 wav eguide standard defi- nition table, offset loss of zero ohm/sec is entered for all f our standards. Lower/minimum frequency Lo wer frequency defines the minimum frequency at which the st andard is to be used f or the purposes of calibration.
16 Upper/maximum frequency This specifies the maximum frequency at which the standard is valid. In broadband applications, a set of banded standards ma y be necessar y to pr o- vide constant response. For example, coaxial offset standards (i.e., 1 / 4 λ offset short) are generally spec- ified o ver bandwidths of an octav e or less.
17 Note Mathematical operations on measurements (and display ed data) after calibration are not corrected for dispersion. Enter W A VEGUIDE int o t he st andar d definition table for all f our standards. Standard labels Labels are entered through the title menu and may contain up to 10 characters.
18 S 11 A,B,C and S 22 A,B,C S 11 A, B,C and S 22 A,B,C cor respond to the S 11 and S 22 ref lection calibrations for por t 1 and port 2 respectivel y. These three classes are used by the Agilent 8510 t o solve for t he systematic errors; directivity , source mat ch, and ref lection tracking.
19 TRL Thru TRL Thru corresponds t o t he measurement of t he S-parameters of a zero-length or shor t t hru connec- tion betw een port 1 and por t 2. The Thr u, Ref lect and Line classes are used ex clusively for t he t hree steps of t he TRL 2-PORT calibration.
20 Each adapter is specified as a sing le delay/thru standard and up to sev en standards numbers can be specified into t he adapt er class. Standard Class labels Standard Class labels are entered t o facilit ate menu-driven calibration.
21 Again, cal kit labels should be chosen t o best describe the calibration devices. The “B.1” default suff ix cor responds t o the kit’s mechanical revision (B) and mathematical revision (1).
22 User modified cal kits and Agilent 8510 specifications As noted previously , the resultant accuracy of the 8510 when used with an y calibration kit is depend- ent on how well its standards are def ined and is verified through measurement of a device wit h traceable frequency response.
23 Appendix A Calibration kit entry procedure Calibration kit specif ications can be ent ered into the Agilent 8510 using the 8510 disk drive, a disk drive connected to the system bus, by front panel entr y , or t hr ough program contr ol by an external controller .
24 Front panel procedure: (P -band waveguide example) 1. Prior to modifying or generating a cal kit, store one or both of the cal kits in t he 8510’s non- volatile memor y t o a disk. 2. Select CAL menu, MORE. 3. Prepare to modify cal kit 2: press MODIFY 2.
25 4. Change the class label for S 11 A: L ABEL CL ASS, S 11 A, ERASE TITLE. 5. Enter t he label of PSHORT 1 b y using the knob, the SELECT sof t key and the SP A CE soft ke y.
26 Appendix B Dimensional considerations in coaxial connectors This appendix describes dimensional considera- tions and required conv entions used in deter min- ing the physical offset lengt h of calibration standards in sexed coaxial connector families.
27 Female type-N Male type-N 7 mm Coaxial connector T ype-N coaxial connector interface The location of the “calibration plane” in T ype-N standards is the outer conductor mating surfaces as shown below . Note: 1.0mm, 1.85mm and 2.4mm connectors not shown, but similar to 3.
28.
29 Appendix C Cal coefficients model Offset devices like of fset shor ts and offset opens can be modeled by t he follo wing signal f low g raph : Figure 1 Signal flow graph model of offset devices The offset portion of t he open or shor t, is modeled as a per f ectly unif orm lossy air dielectric transmis- sion line.
30 Their first or der appro ximations, R is small and G=0, are: Equation 3 Since Equation 4 F or coaxial devices.
31 then: Equation 5 Equation 6 If the Of fset delay=0, then t he coef ficient of ref lection, Γ = Γ L ..
Agilent T echnologies’ T est and Measurement Support, Services, and Assistance Agilent T echnologies aims to maximize t he value y ou receive, while minimizing your risk and problems. W e striv e to ensure that y ou get t he test and measurement capabilities you paid f or and obt ain t he support you need.
An important point after buying a device Agilent Technologies 8510-58 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Agilent Technologies 8510-58 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Agilent Technologies 8510-58 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Agilent Technologies 8510-58 you will learn all the available features of the product, as well as information on its operation. The information that you get Agilent Technologies 8510-58 will certainly help you make a decision on the purchase.
If you already are a holder of Agilent Technologies 8510-58, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Agilent Technologies 8510-58.
However, one of the most important roles played by the user manual is to help in solving problems with Agilent Technologies 8510-58. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Agilent Technologies 8510-58 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center