Instruction/ maintenance manual of the product 66000A Agilent Technologies
Go to page of 26
INSTALLATION GUIDE Agilent Technologies MPS Mainframe Model 66000A SERIAL NUMBERS For mainframe with serial numbers: Agilent Model 66000A: 3138A-00101 and above * For instruments with higher Serial Numbers, a change page may be included. Agilent Part No.
2 CERTIFICATION Agilent Technologies certifies that this product met its published specifications at time of shipment from the factory. Agilent Technologies further certifies that its calibration meas.
3 SAFETY SUMMARY The following general safety precautions must be observed during all phases of operation, service, and repair of this mainframe. Failure to com ply with these precautions or with specific warnings elsewhere in this guide violates safety standards of design, manufacture, and intended use of the mainfram e.
4 S AFETY S UMMARY (continued) GENERAL: Any LEDs used in this product are Class 1 LEDs as per IEC 825-1. ENVIRONMENTAL CONDITIONS: This instrument is intended for indoor use in an installation category II, pollution degree 2 environment. It is designed to operate at a maximum relative humidity of 95% and at altitudes of up to 2000 meters.
5 Printing History The current edition of this guide is indicated below. Reprints of this guide containing minor corrections and updates may have the same printing date. New editions are identified by a new printing date and, in some cases, by a new part number.
6 Table Of Contents 1. General Information Introduction ................................................................................................................ .................................. 9 Safety Considerations .......................
7 Figures 2-1. Power Cord Plug Configurations ............................................................................................ 1 4 2-2. Power Cord Installation ...............................................................................
General Information 9 1 General Information Introduction This guide describes how to install the Agilent 66000A Modular Power System (MPS) Mainframe. It includes the following information: ■ specifi.
General Information 10 Options and Parts The following options are available with the mainframe: Option Description 908 Rack Mounting Kit 909 Rack Mounting Kit with handles 910 Extra Manual Set Power Cords 831 12 AWG unterminated power cord 833 1.
General Information 11 Specifications and Supplemental Characteristics Table 1-1 lists the specifications an d supplemental characteristics of the MPS main fram e. Module specifications are included in the appr o priate module User’ s Guide. Specif ications are warran ted in the temperatu re range of 0 to 55 degrees C.
General Information 12 Table 1-1. Specifications and Supplemental Characteristics (continued) INH/FLT Characteristics Maximum Ratings: ± 16.5 V between pins 3 and 4 + 16.5 V between pins 1 and 2 - 5.3 V between pins 1 and 2 INH (10k pull-up): I I L @ 4 V = 700 µ A I IH @ 5.
Installation 13 2 Installation Installation When you receive your mainframe, first look for any obvious damage that could have occurred during shipment. If there is damage, notify the carrier and the nearest Agilent Sales Office immediately. Warranty information is printed behind the cover page of this manual.
Installation 14 Fuses The mainframe fuse is located inside the mainframe. Refer to the mainframe Service Manual for information about replacing the fuse. Power Cord Installation The power cord supplied with your mainframe may or may not include a power plug.
Installation 15 Figure 2-2. Power Cord Installation Location and Cooling Bench Operation Chapter 1 gives the dimensions of your mainframe. The mainframe has plastic feet that are shaped to ensure self-alignment when stacked with other Agilent System II cabinets.
Installation 16 Rear Panel Switches The six-position rear panel switch is shown in the following figure. Figure 2-3. Rear Panel Switches Address Switches Address switches A0 through A4 set the primary GPIB address. The address is set in binary, A0 being the least significant bit and A4 being the most significant bit.
Installation 17 Mainframe Slot Locations/Addresses Mainframe Mode Switch = MAIN Mainframe Mode Switch = AUX Slot Location 0 1 2 3 4 5 6 7 0 1 2 3 4 5 6 7 Secondary Address 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Note Secondary addresses are reserved for the mainframe whether modules occupy the associated slot locations or not.
Installation 18 Module Installation Note A f ully loaded m ainfram e can weigh over 36 kg (80 lbs). Install the m ainframe into th e rack before installing the modules into the mainf rame. All modules install through the front of the mainframe. Installation (and removal) can be accomplished even when the mainframe is rack-mounted.
Installation 19 ■ There must be no or minimal current flowing through the connector pins at the back of the module that is being removed. This can be accomplished by disabling the module. If this precaution is repeatedly violated at high current, the connector pins at the back of the module will be damaged.
Connections 21 3 Connections Rear Panel Figure 3-1 shows where all of the rear panel connections to the mainframe are made. Figure 3-1. Rear Panel Connections Module Connections Wire connections to the modules are made using a separate module connector that is supplied with each module.
Connections 22 Controller Connections If you have only one MPS mainframe in your system, simply connect it to the controller using the GPIB cable. If you have more than one mainframe, or if you also h.
Connections 23 Figure 3-2. Controller Connections.
Connections 24 Figure 3-3. Digital Connector FLT Output Used to indicate that a fault has occurred on one of the modules in the mainframe. Pins 1 and 2 (pins 1 and 2) are the open-collector output of an optocoupler, with pin 1 the collector and pin 2 the emitter.
Connections 25 Figure 3-4. Examples of the Remote Inhibit Function.
Index 27 Index AM addresses, 16 MAIN, 16, 22 address switch, 16 manuals, 9, 13 Agilent 603xA, 10, 11, 16, 22 module, 11, 21, 22 Agilent 664xA, 10, 11, 16, 22 module - addresses, 22 Agilent 665xA, 10, .
Agilent Sales and Support Offices For more information about Agilent Technologies test and measurement products, applications, services, and for a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir You can also contact one of the following centers and ask for a test and measurement sales representative.
Manual Updates The following updates have been made to this manual since the print revision indicated on the title page. 4/15/00 All references to HP have been changed to Agilent.
An important point after buying a device Agilent Technologies 66000A (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Agilent Technologies 66000A yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Agilent Technologies 66000A - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Agilent Technologies 66000A you will learn all the available features of the product, as well as information on its operation. The information that you get Agilent Technologies 66000A will certainly help you make a decision on the purchase.
If you already are a holder of Agilent Technologies 66000A, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Agilent Technologies 66000A.
However, one of the most important roles played by the user manual is to help in solving problems with Agilent Technologies 66000A. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Agilent Technologies 66000A along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center