Agilent 1604 8A T est Leads Operation and Service Manual Fourth Edition Agilent T echnologi es Japan, Ltd. Agilent PN 160 48-9000 1 March 2 000 Print ed in: Japan.
2 Notices The in format ion con tained in this docume nt is subjec t to ch ange withou t n otice. This docu ment contains p ropri etary infor mation that is p rotec ted by c opyright.
3 • Ground The Instr ument T o avoid electri c shock hazard, t he inst rument chassi s and cabi net must be connecte d to a s afety ea rth gr ound by t he suppli ed power cable with ea rth blade. • DO NOT Operate In An E xplosi ve Atmosphere Do not oper ate the i nstr ument in the prese nce of fl ammable ga sses o r fumes.
4 Wa r r a n t y This Ag ilent T echn ologie s inst rument product is war rante d against defec ts in materia l and workmans hip fo r a perio d corre sponding t o the indi vidual warranty periods of it s compone nt product s. Inst rument s are warr anted for a p eriod of o ne year .
5 Assistance Product ma intena nce agree ments an d other cus t omer a ssistan ce agreements are availa ble for Agilent T echnologi es produc ts. For any as sistance, conta ct your nearest Agilen t T echnologi es Sal es and Servi c e Off ice. Addres ses ar e provide d at t he back of t his manual .
6.
Contents 7 1. Operation Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8 Contents.
9 1 Operation This oper ating no te prov ides co mplete i nformat ion on th e 16048A T est Le ads. The 16048A is s hown pic toriall y in Figur e 1-1, i ts physic al di mensions ar e given in T able 1-1, a nd typi cal char acteristics related to of fset error are give n in T able 1-2.
10 Chapter 1 Operation Product Des cription Product Description The 16048A co nsists of a dir ect att achment, 4-terminal pair type fixture which is equip ped w ith f our BNC (m) termi nat ed-co axia l tes t lead s. Th ese te st lead s ar e used to a ttach use r -fabri cated test fi xture s.
Chapter 1 11 Oper ation Specifications Specifications T able 1-1 Specifications of the 16 048A Function: For use with 4 terminal-pair LCR Meters and Impedance Analy zers. Connector T ype: BNC male Cable Length: Approximately 1m Maximum V oltag e: ± 40V peak max.
12 Chapter 1 Operation T ypica l Character istics T ypical Characteristics T able 1 -2 T ypical Ch aracteristic s NOTE f is the meas urem ent fr eque ncy in meg ahert z. The incr emen tal er rors c alcul ated from the equation i n the t able for measur ements at f requenci es above 1 MHz are additi ve.
Chapter 1 13 Oper ation Compensation for Fixtur e Residual Impedance Error Compensation for Fixtur e R esidual Impedance Err or The 16048A ha s inherent stray c apacitan ce, residual in ductanc e, and res idual resist ance that af fect the accura cy of mea sured val ues.
14 Chapter 1 Operation Compensation for Fixture Residual Impedance Er ror.
15 2 Service.
16 Chapter 2 Serv ice Maintena nce Maintenance An explode d view of the 16 048A (for pa rts i dentifi catio n) is shown in Figure 2-1. Do not dis assemble any furt her than s hown. Maintenance consist s prin cipally o f cleani ng conta cts and r eplacing w orn or d amaged par ts.
Chapter 2 17 Service Maintenance Figur e 2-1 Parts Iden tificat ion.
18 Chapter 2 Serv ice Maintena nce T able 2-1 Parts Id entification Reference Agilent Part No. Qty . Description Note 1 * 4 SLEEVE 2 * 4 NUT 3 * 4 NUT 4 * 4 W ASHER 5 * 4 W ASHER 6 2360-0 192 2 SCREW .
Chapter 2 19 Service Maintenance *: Not se parat ely repla ceable. 31 1250-01 18 4 CONNECTOR-BNC FEMALE 32 236 0-01 15 2 SCREW 1 thru 30 33 160 32-10021 1 P LA TE 34 219 0-0016 4 W ASHER 35 295 0-0001 4 NUT 36 160 32-10022 1 P LA TE 37 160 32-600 01 1 BNC BRA CKET 31thr u36 a.
20 Chapter 2 Serv ice Maintena nce.
REGIONAL SALES AND SUPPORT OFFICES For more information about Agilent Technologies test and measurement products, applications, services, and for a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir. You can also contact one of the following centers and ask for a test and measurement sales representative.
An important point after buying a device Agilent Technologies 16048A (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Agilent Technologies 16048A yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Agilent Technologies 16048A - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Agilent Technologies 16048A you will learn all the available features of the product, as well as information on its operation. The information that you get Agilent Technologies 16048A will certainly help you make a decision on the purchase.
If you already are a holder of Agilent Technologies 16048A, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Agilent Technologies 16048A.
However, one of the most important roles played by the user manual is to help in solving problems with Agilent Technologies 16048A. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Agilent Technologies 16048A along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center